Honeywell Network Card 9782 Series User Manual

9782 Series Two Cell  
Conductivity/Resistivity  
Analyzer/Controller  
Operator’s Manual  
70-82-25-74  
Rev 3  
6/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
About This Document  
Abstract  
This manual contains instructions for installation and operation of the 9782 Series  
Conductivity/Resistivity Analyzer/Controller.  
Revision Notes  
The following list provides notes concerning all revisions of this document.  
Rev. ID  
Date  
Notes  
0
6/98  
This manual accompanies the initial release of the 9782 Series  
Conductivity/Resistivity Analyzer/Controller.  
1
2
3
7/98  
12/98  
6/99  
Changes were made to Section 9 to reflect changes in the software.  
Included dimension drawings for backlit and non-backlit option (Sec. 3)  
Edits were made to standardize terminology and to add the latest MSG.  
References  
Honeywell Documents  
The following list identifies all Honeywell documents that may be sources of reference for the material  
discussed in this publication.  
Document Title  
ID #  
Binder Title  
Binder ID #  
Trademarks  
AutoClean is a trademark of Honeywell Inc.  
Noryl is a trademark of GE Company.  
Contacts  
The following list identifies important contacts within Honeywell.  
Organization  
Telephone  
Address  
Honeywell Technical Assistance Center  
1-800-423-9883  
(USA and Canada)  
1100 Virginia Avenue  
Fort Washington, PA 19038  
6/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
1. INTRODUCTION ..................................................................................................... 1-1  
1.1 Description ..................................................................................................................................... 1-1  
Multi-function instrument...............................................................................................................................1-1  
User interface..................................................................................................................................................1-1  
Easy to configure ............................................................................................................................................1-1  
Input................................................................................................................................................................1-2  
Outputs............................................................................................................................................................1-2  
DE (Digital integration)..................................................................................................................................1-2  
Relays .............................................................................................................................................................1-2  
1.2 Features .......................................................................................................................................... 1-3  
Automatic cell washing...................................................................................................................................1-3  
Computed values.............................................................................................................................................1-3  
Automatic range scaling..................................................................................................................................1-3  
Password protection........................................................................................................................................1-3  
Extensive diagnostics......................................................................................................................................1-3  
Watertight corrosion-resistant case.................................................................................................................1-4  
Solution Temperature Compensation..............................................................................................................1-4  
1.3 Operating the Analyzer/Controller................................................................................................. 1-6  
1.3.1 Keypad Use ........................................................................................................................... 1-6  
Front panel keys used for all operator tasks....................................................................................................1-6  
Selecting a parameter for edit .........................................................................................................................1-7  
Editing a parameter assignment from a list of available choices.....................................................................1-7  
Editing a parameter numeric value .................................................................................................................1-7  
Controlling an AutoClean operation manually................................................................................................1-7  
1.3.2 Online Displays..................................................................................................................... 1-8  
Introduction ....................................................................................................................................................1-8  
Individual cell display.....................................................................................................................................1-8  
Combined display ...........................................................................................................................................1-9  
Computed value..............................................................................................................................................1-9  
Clock and AutoClean....................................................................................................................................1-10  
1.4 Menu Hierarchy............................................................................................................................ 1-11  
Menus for every task.....................................................................................................................................1-11  
1.5 Planning........................................................................................................................................ 1-13  
1.5.1 Overview ............................................................................................................................. 1-13  
Configuration system minimizes decisions...................................................................................................1-13  
1.5.2 Using Relays........................................................................................................................ 1-13  
Relays assigned during I/O setup..................................................................................................................1-13  
Interactions affecting permitted relay assignments .......................................................................................1-14  
1.5.3 Deciding on a Control Strategy........................................................................................... 1-16  
All 9782 models can use relays for control...................................................................................................1-16  
Optional analog outputs for control ..............................................................................................................1-16  
Selecting a strategy.......................................................................................................................................1-16  
1.6 Overview of Installation and Setup Tasks.................................................................................... 1-17  
Setup tasks described in this manual.............................................................................................................1-17  
iv  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. SPECIFICATIONS AND MODEL NUMBER............................................................ 2-1  
2.1 Specifications ................................................................................................................................. 2-1  
2.1.1 User Interface ........................................................................................................................ 2-1  
Display............................................................................................................................................................2-1  
Keypad............................................................................................................................................................2-1  
2.1.2 Physical ................................................................................................................................. 2-1  
Case ................................................................................................................................................................2-1  
Case Dimensions.............................................................................................................................................2-1  
Weight ............................................................................................................................................................2-1  
Mounting ........................................................................................................................................................2-1  
2.1.3 Compliance............................................................................................................................ 2-2  
CE conformity.................................................................................................................................................2-2  
Year 2000 .......................................................................................................................................................2-2  
2.1.4 Electrical................................................................................................................................ 2-2  
Power Requirements.......................................................................................................................................2-2  
Fuse Rating.....................................................................................................................................................2-2  
2.1.5 Model Ranges........................................................................................................................ 2-2  
Standard range models....................................................................................................................................2-2  
Wide range models .........................................................................................................................................2-3  
2.1.6 Operating Conditions ............................................................................................................ 2-3  
Operating Conditions......................................................................................................................................2-3  
Performance under reference operating conditions.........................................................................................2-4  
Operating Influences under normal operating conditions ...............................................................................2-4  
2.1.7 Outputs .................................................................................................................................. 2-4  
Standard alarm/control relays .........................................................................................................................2-4  
Optional alarm/control relays .........................................................................................................................2-4  
Optional analog outputs..................................................................................................................................2-4  
Optional Digital Communication Link............................................................................................................2-5  
2.1.8 Temperature Compensation .................................................................................................. 2-5  
Availability .....................................................................................................................................................2-5  
2.1.9 AutoClean.............................................................................................................................. 2-5  
Cell rinse schedule settings.............................................................................................................................2-5  
2.1.10 Control................................................................................................................................. 2-5  
Settings ...........................................................................................................................................................2-5  
2.1.11 Computed Values ................................................................................................................ 2-6  
Available selections........................................................................................................................................2-6  
2.2 Model Number Breakdown ............................................................................................................ 2-7  
Introduction ....................................................................................................................................................2-7  
3. UNPACKING, PREPARATION, AND MOUNTING ................................................. 3-1  
3.1 Overview ........................................................................................................................................ 3-1  
Introduction ....................................................................................................................................................3-1  
What’s in this section?....................................................................................................................................3-1  
3.2 Unpacking and Preparing ............................................................................................................... 3-2  
Procedure........................................................................................................................................................3-2  
3.3 Mounting ........................................................................................................................................ 3-3  
Introduction ....................................................................................................................................................3-3  
Panel-mounting...............................................................................................................................................3-3  
Wall- and pipe-mounting ................................................................................................................................3-3  
6/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. POWER WIRING..................................................................................................... 4-1  
4.1 Overview ........................................................................................................................................ 4-1  
Introduction ....................................................................................................................................................4-1  
What’s in this section?....................................................................................................................................4-1  
4.2 General Wiring Practices for Power Wiring .................................................................................. 4-2  
Safety precautions...........................................................................................................................................4-2  
Avoid damage to components.........................................................................................................................4-2  
Wiring for immunity compliance....................................................................................................................4-2  
Conform to code .............................................................................................................................................4-3  
4.3 Power Wiring Considerations ........................................................................................................ 4-3  
Recommended wire size .................................................................................................................................4-3  
Power supply voltage and frequency within specs..........................................................................................4-3  
4.4 Installing Power Wiring ................................................................................................................. 4-4  
Procedure........................................................................................................................................................4-4  
5. I/O SETUP AND SYSTEM CONFIGURATION........................................................ 5-1  
5.1 Overview ........................................................................................................................................ 5-1  
Introduction ....................................................................................................................................................5-1  
What’s in this section?....................................................................................................................................5-1  
Powering up the Analyzer/Controller .............................................................................................................5-1  
5.2 I/O Setup and Configuration Tasks................................................................................................ 5-2  
Ease of configuration......................................................................................................................................5-2  
Accessing the I/O setup and configuration menus ..........................................................................................5-2  
Purpose of I/O setup .......................................................................................................................................5-2  
Purpose of configuration menu items..............................................................................................................5-2  
Illustrations .....................................................................................................................................................5-3  
System defaults...............................................................................................................................................5-3  
5.3 Performing I/O Setup ..................................................................................................................... 5-4  
Purpose ...........................................................................................................................................................5-4  
Display resolution and unit of measure...........................................................................................................5-4  
5.4 Configuring Current Adjusting Type (CAT) Control and/or Retransmission of Process Variables5-10  
Purpose .........................................................................................................................................................5-10  
Procedure......................................................................................................................................................5-10  
5.5 Configuring On/Off, Duration Adjusting Type (DAT), or Pulse Frequency Type (PFT) Control5-12  
Purpose .........................................................................................................................................................5-12  
Procedure......................................................................................................................................................5-12  
5.6 Configuring Alarms...................................................................................................................... 5-16  
Purpose .........................................................................................................................................................5-16  
Procedure......................................................................................................................................................5-16  
5.7 Setting the Clock .......................................................................................................................... 5-17  
Purpose .........................................................................................................................................................5-17  
Procedure......................................................................................................................................................5-17  
5.8 Configuring AutoClean ................................................................................................................ 5-18  
Purpose .........................................................................................................................................................5-18  
Timer-initiated operation..............................................................................................................................5-18  
Operator-initiated operation..........................................................................................................................5-18  
Hold mode ....................................................................................................................................................5-19  
vi  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.9 Selecting a Computed Value to Be Displayed ............................................................................. 5-21  
Purpose .........................................................................................................................................................5-21  
Procedure......................................................................................................................................................5-21  
5.10 Summary of Menu Choices and System Defaults...................................................................... 5-23  
Introduction ..................................................................................................................................................5-23  
6. INPUT AND OUTPUT WIRING ............................................................................... 6-1  
6.1 Overview ........................................................................................................................................ 6-1  
Introduction ....................................................................................................................................................6-1  
What’s in this section?....................................................................................................................................6-1  
6.2 General Wiring Practices for I/O Wiring ....................................................................................... 6-2  
Safety precaution ............................................................................................................................................6-2  
Avoid damage to components.........................................................................................................................6-2  
Immunity Compliance.....................................................................................................................................6-2  
Conform to code .............................................................................................................................................6-2  
Recommended maximum wire size.................................................................................................................6-3  
Cables for locations without interference........................................................................................................6-3  
Shielded wiring for locations with interference ..............................................................................................6-3  
Avoiding interference .....................................................................................................................................6-3  
References ......................................................................................................................................................6-3  
6.3 Wiring Analog Inputs and Outputs................................................................................................. 6-4  
Introduction ....................................................................................................................................................6-4  
Cell identification significant to computed value............................................................................................6-4  
Procedure........................................................................................................................................................6-4  
6.4 Wiring Relays................................................................................................................................. 6-7  
6.4.1 Installation............................................................................................................................. 6-7  
Introduction ....................................................................................................................................................6-7  
Procedure........................................................................................................................................................6-7  
6.4.2 Disconnecting RC Arc Suppression Circuits ........................................................................ 6-9  
Introduction ....................................................................................................................................................6-9  
Procedure........................................................................................................................................................6-9  
6.4.3 Wiring a Device to the Relay Indicating Output Range Status........................................... 6-10  
Introduction ..................................................................................................................................................6-10  
Interconnections............................................................................................................................................6-10  
7. AUTOCLEAN THEORY AND PIPING..................................................................... 7-1  
7.1 Overview ........................................................................................................................................ 7-1  
Introduction ....................................................................................................................................................7-1  
What’s in this section?....................................................................................................................................7-1  
7.2 AutoClean Sequence and Piping.................................................................................................... 7-2  
Rinse sequence................................................................................................................................................7-2  
Select valves and fittings with appropriate pressure ratings............................................................................7-3  
Minimize liquid volume in system..................................................................................................................7-3  
6/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. CALIBRATION ........................................................................................................ 8-1  
8.1 Overview ........................................................................................................................................ 8-1  
Introduction ....................................................................................................................................................8-1  
What’s in this section?....................................................................................................................................8-1  
Calibration menu customized for different configurations..............................................................................8-1  
8.2 Recommendations for Successful Measurement and Calibration.................................................. 8-3  
Selection and care of cell essential .................................................................................................................8-3  
Recommendations...........................................................................................................................................8-3  
8.3 Entering the Cal Factor for Each Cell ............................................................................................ 8-4  
Introduction ....................................................................................................................................................8-4  
Procedure........................................................................................................................................................8-4  
8.4 Performing Calibration Trim.......................................................................................................... 8-5  
Introduction ....................................................................................................................................................8-5  
8.5 Removing Calibration Trim Values ............................................................................................... 8-7  
Introduction ....................................................................................................................................................8-7  
Procedure........................................................................................................................................................8-7  
8.6 Determining and Entering the TDS Conversion Factor ................................................................. 8-8  
Introduction ....................................................................................................................................................8-8  
Determining TDS conversion factor...............................................................................................................8-8  
Out-of range-values forced to closest limit.....................................................................................................8-8  
Calibrate the Analyzer/Controller before entering TDS conversion factor.....................................................8-8  
Procedure........................................................................................................................................................8-9  
9. DIAGNOSTICS AND MESSAGES .......................................................................... 9-1  
9.1 Overview ........................................................................................................................................ 9-1  
Introduction ....................................................................................................................................................9-1  
What’s in this section?....................................................................................................................................9-1  
9.2 System Status Messages................................................................................................................. 9-2  
Related to operation of Analyzer/Controller...................................................................................................9-2  
9.3 Process Alarm Messages................................................................................................................ 9-3  
Alarms dependent on configuration................................................................................................................9-3  
9.4 On-Line Diagnostics and System Error Messages ......................................................................... 9-4  
Self-tests at 0.5 second intervals.....................................................................................................................9-4  
Output affected by error..................................................................................................................................9-4  
9.5 Off-Line Diagnostics...................................................................................................................... 9-5  
Introduction ....................................................................................................................................................9-5  
9.5.1 Display and Keyboard Test................................................................................................... 9-5  
Procedure........................................................................................................................................................9-5  
9.5.2 Output Tests .......................................................................................................................... 9-7  
Procedure........................................................................................................................................................9-7  
viii  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10. MAINTENANCE................................................................................................... 10-1  
10.1 Overview .................................................................................................................................... 10-1  
Introduction ..................................................................................................................................................10-1  
What’s in this section?..................................................................................................................................10-1  
10.2 Output Calibration...................................................................................................................... 10-3  
Introduction ..................................................................................................................................................10-3  
Required equipment......................................................................................................................................10-3  
Procedure......................................................................................................................................................10-3  
10.3 Specifying a Tagname or Other Display String ......................................................................... 10-6  
Introduction ..................................................................................................................................................10-6  
Procedure......................................................................................................................................................10-6  
10.4 Viewing Product Information and Changing Model Number Stored in Memory...................... 10-7  
Introduction ..................................................................................................................................................10-7  
Procedure......................................................................................................................................................10-7  
10.5 Adjusting the Screen Contrast.................................................................................................... 10-9  
Introduction ..................................................................................................................................................10-9  
Procedure......................................................................................................................................................10-9  
10.6 Entering a Password for Security............................................................................................. 10-10  
Introduction ................................................................................................................................................10-10  
Procedure....................................................................................................................................................10-10  
10.7 Resetting All Configuration and Calibration Values to Factory Settings................................ 10-11  
Introduction ................................................................................................................................................10-11  
Procedure....................................................................................................................................................10-11  
10.8 Entering Values for Lead Resistance Compensation (Wide Range Only)............................... 10-12  
Introduction ................................................................................................................................................10-12  
Procedure....................................................................................................................................................10-14  
10.9 Replacing the Fuse ................................................................................................................... 10-15  
Introduction ................................................................................................................................................10-15  
Procedure....................................................................................................................................................10-15  
11. ACCESSORIES AND REPLACEMENT PARTS LIST ........................................ 11-1  
11.1 Overview .................................................................................................................................... 11-1  
Introduction ..................................................................................................................................................11-1  
What’s in this section?..................................................................................................................................11-1  
11.2 Part Numbers.............................................................................................................................. 11-2  
Introduction ..................................................................................................................................................11-2  
APPENDIX A PROPORTIONAL CONTROL TUNING ................................................A-1  
Adjust effective proportional band width ......................................................................................................A-1  
Treatment takes too long, but no unacceptable overshoot .............................................................................A-1  
Unacceptable overshoot.................................................................................................................................A-1  
6/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
ix  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX B TWO-CELL APPLICATIONS................................................................B-1  
Ion Exchange .................................................................................................................................................B-1  
Reverse Osmosis............................................................................................................................................B-1  
Conductivity/Resistivity/TDS Difference ......................................................................................................B-2  
Parts Rinsing..................................................................................................................................................B-2  
Softener Monitoring.......................................................................................................................................B-3  
Steam Power Measurements ..........................................................................................................................B-3  
Sodium Hydroxide & Hydrochloric Acid Concentration Measurements.......................................................B-4  
APPENDIX C MICROPROCESSOR BOARD SWITCH SETTINGS ...........................C-1  
Introduction ...................................................................................................................................................C-1  
Location of switches ......................................................................................................................................C-1  
SW1 - all positions set to off .........................................................................................................................C-1  
SW2 - settings are dependent on options.......................................................................................................C-1  
APPENDIX D USING A PRECISION CHECK RESISTOR..........................................D-1  
Introduction ...................................................................................................................................................D-1  
Set cal factor and calibration trim for ideal conditions..................................................................................D-1  
Calculations for conductivity, resistivity, and TDS .......................................................................................D-1  
Concentration values......................................................................................................................................D-2  
x
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Tables  
Table 1-1 Key Functions.......................................................................................................................... 1-6  
Table 1-2 9782 Conductivity/Resistivity Analyzer/Controller Menu Hierarchy................................... 1-11  
Table 1-3 Available Combinations of Relay Assigments ...................................................................... 1-15  
Table 1-4 Overview of Analyzer/Controller Installation Tasks............................................................. 1-17  
Table 3-1 Procedure for Unpacking and Preparing the 9782................................................................... 3-2  
Table 4-1 Procedure for Installing AC Power Wiring ............................................................................. 4-4  
Table 5-1 Procedure for Performing I/O Setup........................................................................................ 5-6  
Table 5-2 Procedure for Configuring CAT Control and/or Retransmission of Process Variables........ 5-10  
Table 5-3 Procedure for Configuring On/Off Control ........................................................................... 5-13  
Table 5-4 Procedure for Configuring PFT Control................................................................................ 5-14  
Table 5-5 Procedure for Configuring DAT Control............................................................................... 5-15  
Table 5-6 Procedure for Configuring Alarms ........................................................................................ 5-16  
Table 5-7 Procedure for Setting the Clock............................................................................................. 5-17  
Table 5-8 Procedure for Configuring AutoClean................................................................................... 5-19  
Table 5-9 Procedure for Specifying a Computed Value ........................................................................ 5-21  
Table 5-10 Menu Choices and System Defaults .................................................................................... 5-23  
Table 6-1 Recommended Maximum Wire Size ....................................................................................... 6-3  
Table 6-2 Procedure for Installing Analog I/O Wiring............................................................................ 6-5  
Table 6-3 Relay Contact Maximum Ratings............................................................................................ 6-7  
Table 6-4 Procedure for Wiring Relays ................................................................................................... 6-8  
Table 6-5 Procedure for Disconnecting RC Arc Suppression Circuits.................................................... 6-9  
Table 8-1 Procedure for Entering Cal Factor........................................................................................... 8-4  
Table 8-2 Conductivity of Potassium Chloride Solutions at 25 °C ......................................................... 8-5  
Table 8-3 Procedure for Performing Calibration Trim Using a Reference Solution ............................... 8-6  
Table 8-4 Procedure for Removing Calibration Trim Values.................................................................. 8-7  
Table 8-5 Procedure for Entering TDS Conversion Factor...................................................................... 8-9  
Table 9-1 System Status Message............................................................................................................ 9-2  
Table 9-2 Process Alarm Messages.......................................................................................................... 9-3  
Table 9-3 On-Line Diagnostics and System Error Messages................................................................... 9-4  
Table 9-4 Procedure For Initiating Display and Keyboard Tests............................................................. 9-5  
Table 9-5 Procedure for Testing Relays and Analog Outputs ................................................................. 9-7  
Table 10-1 Procedure for Calibrating Outputs....................................................................................... 10-4  
Table 10-2 Procedure for Specifying a Tagname or Other Display String............................................ 10-6  
Table 10-3 Procedure for Viewing Product Information and Changing the Stored Model Number.... 10-7  
Table 10-4 Procedure for Adjusting the Screen Contrast ...................................................................... 10-9  
Table 10-5 Procedure for Entering a Password.................................................................................... 10-10  
Table 10-6 Procedure for Resetting All Configuration to Factory Settings......................................... 10-11  
Table 10-7 Procedure for Entering Values for Lead Resistance Compensation.................................. 10-14  
6/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
xi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 10-8 Procedure for Replacing the Fuse...................................................................................... 10-16  
Table 11-1 Part Numbers ....................................................................................................................... 11-2  
xii  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figures  
Figure 1-1 9782 Front Panel..................................................................................................................... 1-5  
Figure 1-2 Representative Process Value Display for a Cell in Standard Range Analyzer..................... 1-8  
Figure 1-3 Representative Combined Display ......................................................................................... 1-9  
Figure 1-4 Representative Computed Value Display............................................................................... 1-9  
Figure 1-5 AutoClean Display ............................................................................................................... 1-10  
Figure 3-1 Sample Nameplate.................................................................................................................. 3-3  
Figure 3-2-1 Mounting and Dimensions - Non-Backlit Display Option.................................................. 3-4  
Figure 3-2-2 Mounting and Dimensions - Non-Backlit Display Option.................................................. 3-4  
Figure 3-3-1 Mounting and Dimensions - Backlit Display Option.......................................................... 3-4  
Figure 3-3-2 Mounting and Dimensions - Backlit Display Option.......................................................... 3-4  
Figure 4-1 AC Power Terminals .............................................................................................................. 4-5  
Figure 6-1 Terminal Connections............................................................................................................. 6-6  
Figure 6-2 Sample Multipoint Recorder Chart Illustrating Auto Output Range Change for Steam  
Condensate Monitoring.................................................................................................................... 6-10  
Figure 6-3 Interconnections between 9782 and Recorder with Range Identification by Point Change 6-11  
Figure 7-1 Automatic Cell Wash Setup ................................................................................................... 7-3  
Figure 10-1 Example of a Conductivity Loop...................................................................................... 10-14  
Figure 10-2 Location of Power Fuse.................................................................................................... 10-18  
6/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
xiii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
xiv  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
1. Introduction  
1.1 Description  
Multi-function instrument  
The Honeywell 9782 Conductivity/Resistivity Analyzer/Controller (Figure 1-1) is a  
microprocessor-based instrument for analysis of conductivity, resistivity, total dissolved solids,  
or concentration in industrial processes. The resolution for all measurements is +/- 0.5% of  
reading (except for temperature; see below). The units of measure are:  
conductivity: microSiemens per centimeter, microSiemens per meter, or milliSiemens per  
centimeter  
resistivity: megohms or kilohms  
TDS: PPT, PPM or PPB  
concentration: percent by weight  
temperature: resolution 0.1% for 0 to 99.9 °C, and 1 C° for 100 C° and higher  
User interface  
An easy-to-read display provides instant access to process values. The displays of process values  
show the instrument’s tagname (or other configurable text) at the top of the screen.  
Every display includes an “alarm stripe”, a line dedicated to displaying process alarm messages,  
status messages, and system diagnostic messages, whenever any are active. (One or more relays  
can also be used for alarm annunciation; see 1.5.2.)  
Dedicated-function membrane keys to the right of the display on the front panel are used to:  
cycle through the displays of real-time values  
access configuration, calibration, and maintenance menus  
change and enter setpoints and other parameters  
In addition, three variable function keys are below the display. The function of each of these  
“soft keys” depends on the purpose of the screen currently on display. Soft key labels are always  
displayed at the bottom of the screen, immediately above the relevant function key.  
Easy to configure  
Menu-driven configuration is fast and easy. Only configuration parameters related to features  
supported by the hardware and relevant to the specific application are displayed.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Input  
Input to the 9782 can be from any Honeywell conductivity cell at distances up to 1500 feet.  
Outputs  
Optional isolated 0 to 1 V, 0 to 10 V and 4 to 20 mA outputs are available. Use these analog  
outputs for retransmission of process variables, or for Current Adjusting Type control using an  
output signal that is directly proportional to the input (see 1.5.3).  
Relays  
Two 2A SPDT alarm/control relays are standard, with an additional two 3A relays (either general  
purpose or hermetically sealed) available as an option (see Sub-Section 2.17). These relays can  
be used to:  
take advantage of special features such as automatic cleaning of cells  
control process variables (see 1.5.3)  
annunciate alarms  
More details about use of relays are provided in 1.5.2.  
1-2  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
1.2 Features  
Automatic cell washing  
The standard AutoClean feature can periodically rinse cells automatically. The schedule and  
duration of the operation are configurable. An internal relay actuates external solenoid valves to  
control the flow of rinse water to the cells.  
Computed values  
The Analyzer/Controller can calculate and display your choice of the following values:  
ratio of the two cell inputs  
difference between two cell inputs  
percent passage  
percent rejection  
parts per million carbon dioxide (9782C-S0 only)  
The equations used are in the specifications in Section 2.  
Automatic range scaling  
The Analyzer/Controller supports automatic range scaling of the output. If this feature is  
enabled, then when the measured value falls below 10% of the output scaling limit, the output is  
scaled by a factor of 10. An internal relay can be dedicated to this function. The relay closes  
contacts  
NC-COM when the output is scaled to the low range, and closes contacts NO-COM when the  
range shift occurs. The relay can be used to control an indicator or to switch the output to a  
second channel of a multipoint recorder when the range change occurs.  
This output range scaling has no effect on the display. Display scaling is specified independent  
of output scaling.  
Password protection  
A password (up to four digits) can be configured. If the security feature is enabled, the password  
will be required to access configuration, calibration, and maintenance software functions.  
Extensive diagnostics  
The 9782 Analyzer/Controller performs extensive self-diagnostics as a background task during  
normal operation. If a problem is detected, a message is displayed on the alarm stripe to alert the  
operator. In addition, the operator can initiate keypad and display tests using Maintenance Menu  
functions.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
1-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Watertight corrosion-resistant case  
The 9782 is enclosed in a watertight and corrosion resistant industrial case, designed for panel,  
pipe or wall mounting. It has an EMI/RFI shielded plastic case.  
Solution Temperature Compensation  
There are two types of 9782C Analyzer/Controller models: standard range and wide range. Both  
types provide temperature compensation as described below. The standard range  
Analyzer/Controller is ideal for high-purity water applications. The wide range  
Analyzer/Controller is suitable for chemical solutions.  
To take advantage of the temperature compensation capabilities, the appropriate solution  
temperature compensation type must be entered for each cell as described in Section 5.  
Standard Range - A standard range 9782 Analyzer/Controller provides true high purity  
temperature compensation for deionized water as well as for routine applications. The  
Analyzer/Controller also employs unique compensation algorithms for the highest accuracy in  
power industry water treatment using ammonia or morpholine, as well as for cation conductivity  
samples. Semiconductor resistivity measurements on acid etch rinse operations, especially with  
hot deionized water, also achieve the highest available accuracy. For pharmaceutical  
applications, raw conductivity data is available at the touch of a button.  
For high-purity water special temperature compensation is required in the range of 0.055 to 5  
microsiemens/cm per 0.2 to 18 Megohms-cm. Traditional compensation techniques used in most  
analog instruments require temperature compensator selection for a reference conductivity  
represented by a single curve. This method lacks the capability to accurately reflect the changing  
temperature coefficient of water for variations in temperature and conductivity on either side of  
that reference value. Advanced high-purity water temperature compensation determines the  
conductivity of an impurity at any concentration of water at any temperature, and the interaction  
between the two. It compensates these values to obtain the conductivity referenced to 25 °C.  
Because the temperature coefficients involved are both high and variable, a precise temperature  
measurement is required to achieve accurate compensation. To accomplish this the 9782  
Analyzer/Controller first makes precise temperature measurements over a wide range, even with  
long sensor leads. Second, an accurate conductivity measurement is assured by use of the cell  
constant trim factor. Third, internal high resolution algorithms are used to make exact high-  
purity water calculations. This high level of performance in an inexpensive, yet highly reliable  
instrument is the result of over seventy years of experience in making industrial conductivity  
measurements.  
For power plant cation/ammonia conductivity measurements temperature compensation provides  
the highest accuracy for measurements of cation exchanger effluent and acidic cation  
conductivity samples with Larson Lane-type analyzers.  
This compensation matches both strong inorganic and weak organic acid temperature  
characteristics. Measurements of steam and condensate samples containing ammonia are also  
greatly improved using this algorithm.  
For power plant morphonline conductivity measurements temperature compensation was  
developed specifically for this market’s water chemistry. Morpholine has unique temperature  
characteristics that are handled automatically by the 9782 Analyzer/Controller.  
1-4  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
For semiconductor etch rinse resistivity measurements temperature compensation provides the  
most accurate measurements of used rinsewater containing traces of acidic etchants. Rinse  
quality control and water reclamation monitoring can be made with greater precision when this  
compensation is used.  
Wide Range - A wide range Analyzer/Controller uses precise algorithms to compute  
temperature-compensated conductivity, total dissolved solids (TDS), or concentration for display  
and output, based on conductivity and temperature inputs. The wide range instruments are ideal  
for controlling deionizer and softener regenerant dilutions. Full spectrum temperature  
compensation, specifically for sodium hydroxide, sulfuric acid, hydrochloric acid and sodium  
chloride, automatically provides exact correction of temperature influences as they vary with  
both concentration and temperature. Readout in conductivity or ppm TDS units may also be  
selected.  
F2  
F3  
F1  
+RQH\ZHOO  
Figure 1-1 9782 Front Panel  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
1-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
1.3 Operating the Analyzer/Controller  
1.3.1 Keypad Use  
Front panel keys used for all operator tasks  
As shown in Figure 1-1, five keys with dedicated functions are on the front panel. In addition,  
three “soft keys” vary their function according to the needs of the screen on display. Use of the  
keys is described in Table 1-1.  
Table 1-1 Key Functions  
Key  
Name  
Use  
DISPLAY  
When process values are on display: Use DISPLAY to cycle  
through available real-time displays.  
When the main menu (Configuration, Calibration,  
Maintenance, I/O Setup) is on display: Use DISPLAY to  
return to displaying process values.  
MENU  
When process values are on display: Use MENU to access  
the menus. If the security feature is enabled, the display will  
prompt for entry of the password before access to the menus  
is permitted. (Enter the password using the procedure for  
editing a parameter numeric value as described on the next  
page.)  
When any menu or configuration screen is on display: Use  
MENU to go up a level in the display hierarchy. Frequently this  
means returning to the main menu.  
UP  
When a menu or configuration screen is on display: Use  
UP to highlight a different item.  
When changing a numerical value: Use UP key to increment  
the value of the digit at the cursor.  
DOWN  
ENTER  
When a menu or configuration screen is on display: Use  
DOWN to highlight a different item.  
When changing a numerical value: Use DOWN to  
decrement the value of the digit at the cursor.  
When a menu item is highlighted: Use ENTER to select it.  
When editing a parameter: Use ENTER to save the new  
value.  
F1, F2, F3  
[function  
keys]  
When a “soft key” label is displayed below the alarm  
stripe: Use the function key directly below the label to perform  
the action.  
F1  
1-6  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
Selecting a parameter for edit  
To select a parameter for edit:  
display the screen containing the parameter  
use the UP or DOWN keys to highlight the parameter name  
press the ENTER key to highlight the displayed current value  
Editing a parameter assignment from a list of available choices  
To edit a parameter having a text string as an assigned value:  
select the parameter and highlight its current value as described above  
use the UP or DOWN keys to display the other valid choices  
when your choice is displayed, press ENTER  
Editing a parameter numeric value  
To edit a parameter having a numeric value:  
select the parameter and highlight its current value as described above; the “DPP”, ←  
and soft keys will be displayed (During calibration the and soft keys will be  
displayed once the UP or DOWN key has been pressed.)  
use the “DPP” soft key to position the decimal point  
use the or soft keys to move the cursor to the digit to be changed  
use the UP and DOWN keys to increment or decrement the value  
when all digits necessary have been changed, press ENTER  
Controlling an AutoClean operation manually  
To initiate an AutoClean operation manually, press the DISPLAY key until the AutoClean  
screen is displayed. Next press the “START” soft key.  
This display can also be used to see how much time is remaining until the next scheduled  
AutoClean operation. When AutoClean is active, the alarm stripe will display “AUTO CLEAN”,  
and the AutoClean display will show 00:00:00.  
If necessary, pause the operation using the “PAUSE” soft key. To resume the operation, press  
“PAUSE” again. To cancel the operation, use the “STOP” soft key. If pause is left on, the  
sequence will resume automatically after twenty minutes.  
At the start of the AutoClean operation alarms and outputs will be held at current levels,  
regardless of whether the AutoClean was scheduled or started manually. The alarms and outputs  
will be held throughout the AutoClean operation.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
1-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
1.3.2 Online Displays  
Introduction  
The online displays available when the operator presses the DISPLAY key depend on the  
number of cells connected to the Analyzer/Controller, and on the special features used. The  
instrument’s tagname (or other configurable text) is at the top of the screen for all displays  
except the digital clock and the AutoClean Sequence display. Every display includes an “alarm  
stripe”, a line dedicated to displaying process alarm messages, status messages, and system  
diagnostic messages, whenever any are active.  
Individual cell display  
For each cell connected to the Analyzer/Controller an online display similar to Figure 1-2 is  
provided. The information on this display includes:  
the value measured by the cell, with the appropriate unit of measure filled in  
an indication of whether the value is temperature compensated (TC) or not  
compensated (NC); the display can be toggled between displaying compensated and  
raw data using the F1 function key  
the temperature in °C  
if the value displayed is a percent concentration (9782C-W0 only), the chemical will  
be identified above the alarm stripe  
In addition of the function key used to toggle the temperature compensation display, function  
keys are available to access calibration displays (without going through the menus), and to  
acknowledge alarm stripe messages.  
CONDENSATE  
T105  
TC  
C1  
µS  
8.59  
25.2°C  
ALARM STRIPE  
NC  
CAL  
ACK  
Figure 1-2 Representative Process Value Display for a Cell in Standard Range Analyzer  
1-8  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
Combined display  
If two cells are connected to the Analyzer/Controller, then a display is available that includes the  
measured variable for both cells (see Figure 1-3). Another combined display also includes the  
temperature measured by each cell.  
MIXED BEDS  
1 AND 2  
TC  
C1  
C2  
0.58  
µS  
µS  
0.25  
ALARM STRIPE  
CAL  
NC  
ACK  
Figure 1-3 Representative Combined Display  
Computed value  
If the Analyzer/Controller is configured to calculate a computed value such as the ratio between  
the cells’ readings, a screen will be available to display the value, along with a label (see Figure  
1-4).  
CATION TRAIN A  
0.67  
RATIO  
ALARM STRIPE  
NC  
CAL  
ACK  
Figure 1-4 Representative Computed Value Display  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
1-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Clock and AutoClean  
If the 24-hour clock is turned on, a time display will be available. If AutoClean is configured,  
then the AutoClean screen will be included in the display cycle (see Figure 1-5). When  
AutoClean is active, the screen will display the time remaining. When this screen is on display,  
the function keys can be used to control the AutoClean operation as described in 1.3.1.  
AUTO CLEAN  
TIME REMAINING  
01: 58:29  
ALARM STRIPE  
START  
PAUSE  
STOP  
Figure 1-5 AutoClean Display  
1-10  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
1.4 Menu Hierarchy  
Menus for every task  
For your convenience, menus are provided for configuration, calibration, maintenance, and I/O  
setup tasks. The menu hierarchy is shown in Table 1-2. Not every menu item applies to every  
system.  
Table 1-2 9782 Conductivity/Resistivity Analyzer/Controller Menu Hierarchy  
Menu Item  
Purpose  
Where  
Described  
set up Analyzer/Controller to implement  
your alarm and control strategy  
Section 5  
5.4  
CONFIGURATION  
CAT/RETRANSMISSION  
configure parameters for using analog  
outputs for control and retransmission of  
process values  
ALARMS  
specify value to be alarmed, alarm  
setpoints, deadbands, and delay  
5.6  
5.5  
DISCRETE CONTROL  
ADVANCED FEATURES  
specify setpoints, range limits, etc. to use  
relays for control  
set the clock, set up automatic cell  
cleaning, and specify the computed value  
to be displayed  
5.7, 5.8, 5.9  
calibration functions  
Section 8  
CALIBRATION  
CELL CAL FACTOR  
CALIBRATION TRIM  
enter cell-specific cal factor  
8.3  
calibrate system using a reference solution  
with each cell; remove trim  
8.4, 8.5  
TDS CONVERSION FACTOR C1  
TDS CONVERSION FACTOR C2  
enter a process-specific total dissolved  
solids conversion factor for each cell  
8.6  
Table 1-2 continued on next page  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
1-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Table 1-2 9782 Conductivity/Resistivity Analyzer/Controller Menu Hierarchy  
Menu Item  
Purpose  
Where  
Described  
do occasional system tasks  
Sections 9 and  
10  
MAINTENANCE  
OFF-LINE FUNCTIONS  
DISPLAY TEST  
[see functions listed below]  
test display  
[see below]  
9.5.1  
KEYBOARD TEST  
test keyboard  
9.5.1  
OUTPUT TESTS  
test relays and analog outputs  
calibrate outputs electrically  
[see functions listed below]  
9.5.2  
OUTPUT CALIBRATION  
INSTRUMENT SETUP  
INSTRUMENT TAGGING  
10.2  
[see below]  
10.3  
configure tagname or other text to be  
displayed at the top of the screen  
PRODUCT INFORMATION  
view software version and stored model  
number; if a special technical support  
password is known, alter model number in  
memory (in case of hardware upgrades)  
10.4  
SCREEN CONTRAST  
SECURITY  
adjust display contrast  
10.5  
10.6  
assign password to limit access to  
configuration, calibration, and maintenance  
functions  
RESET UNIT  
reset all configuration and calibration  
values back to factory settings  
10.7  
10.8  
LEAD RESISTANCE COMP  
[wide range models only]  
specify wire gauge and length when long  
leads are used; instrument will compensate  
for these long leads  
choose features, make relay and analog  
output assignments  
5.3  
I/O SETUP  
1-12  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
1.5 Planning  
1.5.1 Overview  
Configuration system minimizes decisions  
The 9782C Analyzer/Controller was designed for easy configuration using menu displays and the  
keys on the front panel. Numeric values such as setpoints and range limits are easy to enter using  
the UP and DOWN keys and function keys (see 1.3). Other configuration parameter  
assignments are selected by scrolling through a list of available choices using the UP and  
DOWN keys.  
During configuration only those parameters and choices relevant to features supported by your  
Analyzer/Controller’s hardware will be available for configuration. For example, if the model  
purchased does not contain hardware for the optional analog outputs, then none of the screens  
relating to output configuration will be applicable.  
Although configuration is a simple process, to make best use of the 9782C Analyzer/Controller  
and its features, a little advance planning is recommended.  
1.5.2 Using Relays  
Relays assigned during I/O setup  
Every 9782 model includes two internal relays. Two more relays are available as an option.  
During I/O setup (see 5.3) the 9782’s software will assign relays to the features you select. The  
design of the Analyzer/Controller allows multiple alarms to be assigned to one relay, allowing  
greater flexibility with fewer relays. All other 9782 features using relays (AutoClean, range  
status indication, and discrete control) require one relay per instance of the feature. You will not  
be permitted to select features requiring more that the available number of relays.  
Once a relay has been dedicated to a function, the relay will not be available for assignment to  
another function. In order to get the most functionality from the available relays, plan your  
strategy before implementation. The order in which features are presented (during I/O setup) for  
relay assignment is:  
1. Cell 1 “Advanced Features” AutoClean and range status indication  
2. discrete control (both cells)  
3. alarm annunciation (both cells), and Cell 2 AutoClean and range status  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
1-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Interactions affecting permitted relay assignments  
Table 1-3 is provided so that you can optimize your use of relays to implement your control  
strategy. Each row in the table represents an available combination of relay assignments for a  
single feature or multiple alarms.  
ATTENTION  
AutoClean and range status indication cannot be selected for use with two cells simultaneously.  
Your choices are:  
For Cell 1 only, use AutoClean (Relay 1) and/or range status indication (Relay 2).  
For both Cell 1 and Cell 2, use AutoClean (Relays 1 and 3), with no range status indication,  
nor any discrete control.  
For both Cell 1 and Cell 2, use range status indication (Relays 2 and 4), with no AutoClean,  
nor any discrete control.  
When reading the tables keep in mind the following:  
The entry “Diagnostic” in the tables indicates that the relay can be used to indicate the  
unit’s failure of a self-test (see 9.4).  
The entry “or discrete control” means that the relay can be assigned to discrete control  
or to one alarm. If the relay is selected for discrete control, then the alarm choice  
sharing the cell in the table with “or discrete control” will not be available in that  
row’s combination of relay assignments.  
If the Analyzer/Controller has only two relays, then only the choices for Relay 1 and  
Relay 2 will be available.  
1-14  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
Table 1-3 Available Combinations of Relay Assigments  
Features Used  
Alarms Only  
Relay 1  
Relay 2  
Relay 3  
Relay 4  
Alarm 1  
Alarm 2  
Alarm 2  
Alarm 3  
Alarm 3  
Alarm 3  
Alarm 3  
Alarm 4  
Alarm 4  
Alarm 4  
Alarm 1 and Diagnostic  
Alarm 1  
Alarm 2 and Diagnostic  
Alarm 2 and Alarm 4  
Alarm 3 and Alarm 4  
Alarm 1  
Alarm 2 and Alarm 4  
Alarm 3 and Alarm 4  
Alarm 1 and Alarm 2  
Alarm 1 and Alarm 2  
Alarm 1 or discrete control Alarm 2 or discrete control Alarm 3 or discrete control Alarm 4 or discrete control  
Alarm 1 and Diagnostic  
Alarm 2 or discrete control Alarm 3 or discrete control Alarm 4 or discrete control  
Discrete Control  
and Alarms  
Alarm 1 or discrete control  
Alarm 2 and Diagnostic  
Alarm 3 or discrete control Alarm 4 or discrete control  
Alarm 3 or discrete control Alarm 2 and Alarm 4  
Alarm 1 or discrete control Alarm 2 and Alarm 4  
Alarm 1  
Cell 1 Range Status  
Cell 1 Range Status  
Cell 1 Range Status  
Alarm 3  
Alarm 3  
Alarm 3  
Alarm 4  
Alarm 4  
Alarm 4  
Alarm 1 and Alarm 2  
Alarm 1 and Diagnostic  
Range Status  
and Alarms  
Alarm 1 and 2 and 3 and 4 Cell 1 Range Status  
Alarm 1 and 2 and 3 and 4 Cell 2 Range Status  
Alarm 1 and 2 and 3 and 4 Cell 1 Range Status  
and Diagnostic  
Alarm 1 and 2 and 3 and 4 Cell 2 Range Status  
and Diagnostic  
Cell 1 Range Status  
Cell 1 Range Status  
Cell 1 Range Status  
Alarm 3 or discrete control Alarm 4 or discrete control  
Alarm 3 or discrete control Alarm 4 or discrete control  
Alarm 3 or discrete control Alarm 4 or discrete control  
Range Status,  
Discrete Control  
and Alarms  
Alarm 1 or discrete control  
Alarm 1 and Alarm 2  
Alarm 1 and Diagnostic  
AutoClean  
AutoClean  
Cell 1 Range Status  
Cell 1 Range Status  
Alarm 3 or discrete control Alarm 4 or discrete control  
AutoClean, Range  
Status, and Alarms  
Alarm 2 and 3 and  
Diagnostic  
Alarm 4 or discrete control  
With or Without  
Discrete Control  
AutoClean  
AutoClean  
AutoClean  
AutoClean  
Alarm 2  
Alarm 3  
Alarm 4  
Alarm 2 and Diagnostic  
Alarm 2 and 3 and 4  
Alarm 3  
Alarm 4  
AutoClean  
and Alarms  
AutoClean  
AutoClean  
Alarm 2 and 3 and 4  
Alarm 2 and 3 and 4 and  
Diagnostic  
Alarm 2 and 3 and 4 and  
Diagnostic  
AutoClean  
AutoClean  
Alarm 2 or discrete control Alarm 3 or discrete control Alarm 4 or discrete control  
Alarm 2 and Diagnostic  
Alarm 3 or discrete control Alarm 4 or discrete control  
AutoClean,  
Discrete Control,  
and Alarms  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
1-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
1.5.3 Deciding on a Control Strategy  
All 9782 models can use relays for control  
All 9782 models have at least two relays. (Two more relays are optional.) Unless dedicated to  
other functions, these relays can be used to control process variables using the three types of  
discrete control available.  
On/Off Control using one to four cycle timers with configurable setpoint, deadband,  
cycle period, and “on” time.  
PFT (Pulse Frequency Type) Proportional-Only Control using one to four relays. The  
pulse output is generated by repeated relay contact opening and closure. The frequency  
of the pulse is proportional to the deviation from the configurable setpoint. The  
proportional band limit and maximum pulse frequency rate are also configurable. This  
type of output is used to control such devices as pulse-type electronic metering pumps.  
DAT (Duration Adjusting Type) Proportional-Only Control using one to four relays.  
Also known as time-proportioned control, DAT cycles a relay output on and off,  
varying the “on” time in proportion to the deviation from setpoint. This type of output  
is used to control devices such as solenoid valves.  
Optional analog outputs for control  
Up to three analog outputs are available as options. These current or voltage outputs can be used  
to retransmit a process variable using an output signal that has been scaled. An analog output  
that is directly proportional to an input can be used for Current Adjusting Type (CAT) control.  
(Despite the name of this control type, voltage outputs can also be used.)  
When an output is used to retransmit a process variable, the values representing 0 and 100%  
outputs are configurable.  
In CAT control the process values for the setpoint and proportional band limit are configurable.  
Selecting a strategy  
When considering whether to use On/Off control or one of the proportional-only strategies,  
compare the requirements of your application with the benefits of each control type.  
Proportional control will usually provide faster batch treatment with less overshoot than On/Off  
control. However, it is important to note that proportional-only control is recommended only for  
batch processes and for pretreatment in continuous neutralization systems.  
Proportional-only control does not include any reset action. As a result, changes in reagent  
demand that occur in continuous processes are not accommodated and control will not reach the  
setpoint. In a batch process, reagent is added until a setpoint is reached and the batch is not  
emptied until this occurs.  
If you decide to use proportional-only control, select the type based on the input requirements of  
the device to be controlled. See Appendix A for a discussion of tuning proportional-only control.  
1-16  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
1.6 Overview of Installation and Setup Tasks  
Setup tasks described in this manual  
This manual contains instructions for all installation and operation tasks relating to the  
Analyzer/Controller. (Instructions for installing and using the cells are provided in the manuals  
supplied with the cells.) Table 1-3 provides an overview of the Analyzer/Controller installation  
tasks, as well as providing references to the relevant sections of the manual.  
Table 1-4 Overview of Analyzer/Controller Installation Tasks  
Sequence  
1
Task  
Section  
Consider the environmental and electrical specs when  
selecting a site to install the Analyzer/Controller.  
2
Unpack, inspect and mount the unit.  
3
4
2
3
Check tag to confirm that unit is compatible with the voltage at  
your site.  
Install power wiring.  
4
5
4
5
Perform I/O setup and configure the software to implement  
your control strategy.  
Wire inputs and outputs to match I/O setup in Step 5.  
6
7
6
7
8
If AutoClean will be used, install necessary piping and valves.  
Configure a tagname or other text string to be displayed at the  
top of the screens showing real-time process values.  
10  
To limit access to I/O setup, configuration, calibration, and  
maintenance functions, enter a password.  
10  
10  
9
Display model number stored in memory and software version  
number. Note these for future reference. (If you call for  
technical assistance, you will need this information.)  
10  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
1-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
The manual also contains:  
information about diagnostics, status messages, and system error messages  
(Section 9)  
instructions for returning all parameter values to the factory settings (Section 10)  
instructions for calibrating the outputs and changing the fuse (Section 10)  
parts list (Section 11)  
supplementary information for special applications and proportional control tuning tips  
(appendices).  
1-18  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Model Number  
2. Specifications and Model Number  
2.1 Specifications  
2.1.1 User Interface  
Display  
LCD dot matrix display, 128 by 64 dpi. Backlit: solid state LED. Displays Cell 1 and Cell 2  
values in terms of conductivity, resistivity, concentration, or TDS (temperature compensated or  
not compensated); Cell 1 and Cell 2 temperature; computed value (if enabled); time (if clock is  
enabled); time remaining in AutoClean cycle (if feature is configuration); alarm conditions, and  
diagnostics.  
Alarm messages, status messages, and diagnostic messages are available on all displays.  
Keypad  
Monoplanar front panel with 8 keys. Push-button entry with tactile feedback.  
2.1.2 Physical  
Case  
Gray Noryl plastic, NEMA 4X/IP55 Rainproof and Outdoor. Interior conductive coating to  
provide effective RFI/EMI shielding.  
Case Dimensions  
156 x 156 x 178 mm (6 1/8 x 6 1/8 x 7 in.); panel cutout 141 x 141 mm (5.53 x 5.53 in.).  
1.8 kg (4 lb).  
Weight  
Mounting  
Panel mounting hardware supplied. Surface and pipe mounting hardware available. See Figure  
3-2 or the Parts List.  
2.1.3 Compliance  
FM/CSA  
This product has FM and CSA Hazardous Location Approval for Class I, Div. 2, Groups A, B, C  
and D classified areas.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
CE conformity  
This product is in conformance with the protection requirements of the following European  
Council Directives: 89/336/EEC, the Electromagnetic Compatibility Directive and 73/23/EEC,  
the Low Voltage Directive. Conformance of this product with any other “CE Mark” Directive(s)  
shall not be assumed.  
Year 2000  
This instrument will be unaffected by the occurrence of the year 2000.  
2.1.4 Electrical  
Power Requirements  
120 +/- 10% Vac, 47-63 Hz, 15 VA.  
240 +/- 10% Vac, 47-63 Hz, 15 VA.  
Memory retained by EEPROM when power is off.  
Fuse Rating  
120V operation : 0.25 amp/250V  
240V operation: 0.125 amp/250V  
2.1.5 Model Ranges  
Standard range models  
Standard Display Ranges: 9782C-S0  
Conductivity  
(microsiemens/cm)  
TDS**  
Resistivity (ohms-cm) Cell Constant Required  
(cm-1)  
0-1.999*  
0-19.99*  
0-199.9*  
0-1,999  
0-1,999 ppb  
1-19,990 ppb  
0-199.9 ppm  
0-1,999 ppm  
0-10,000 ppm  
0-19.99M  
0-1.999M  
0-199.9K  
0-19.99K  
0-1.999k  
0.01  
0.1***  
1.0  
10  
0-19,990  
25 or 50  
* Equivalent µS/m ranges also available.  
** For power plant degassed cation conductivity measurements.  
*** Recommended cell constant for most power plant applications.  
2-2  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Model Number  
Wide range models  
Wide Display Ranges: 9782C-W0  
Conductivity (microsiemens/cm)  
TDS  
Cell Constant Required (cm-1)  
100-1,999  
1000-19,990  
0-1,999 ppm  
0-19,990 ppm  
0.1  
1
Conductivity (millisiemens/cm)  
10-199.9  
0-199.9 ppt  
10  
25  
50  
10-500  
20-1,000  
Concentration  
0-18.00% Hydrochloric Acid*  
0-19.99% Sulfuric Acid  
0-19.99% Sodium Hydroxide*  
0-19.99% Sodium Chloride  
50**  
50**  
25** ro 50  
25** or 50  
*
The maximum concentration measurable by conductivity depends on temperature. Maximum  
for sodium hydroxide allowed by the 9782 Analyzer is 10% at 0º C, 13% at 25º C and 20% at  
75º C with linear interpolation between these points. The maximum concentration for  
hydrochloric acid is 15.5% above 40º C.  
** Measurements in the lower half of the concentration ranges and below 50º C may also use  
the next lower available cell constant.  
2.1.6 Operating Conditions  
Operating Conditions  
Ambient Temperature: Normal 0 to 60 °C (32 to 140 °F), extreme -10 to +60 °C (14 to 140 °F);  
storage -30 to +70°C (-22 to 158 °F).  
Line Voltage: Normal 120 +/- 10% Vac (extreme 100 to 132 Vac, 47 to 63 Hz)  
Normal 240 +/- 10% Vac (extreme 200 to 264 Vac, 50 to 60 Hz)  
RH: 90% maximum non-condensing @ 40°C (104 °F) max.  
Installation Category: II  
Pollution Degree: 2  
Altitude: <20,000 M (6562 ft)  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
2-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Performance under reference operating conditions  
Display Accuracy:  
Standard Ranges (Conductivity/Resistivity); ±0.5% of reading.  
Wide Ranges (Conductivity); ±0.5% of reading or 0.02% of span.  
Concentration; ±0.5% of span, ±1% conformity to conductivity vs. concentration data tables.  
Output accuracy: Adjustable to within 0.01 mA, repeatable to within ±0.1% of span.  
Reference Operating Conditions: 25 ±1 °C; 10 to 40% RH; 120 or 240 Vac; process temp 25 °C  
measured  
Operating Influences under normal operating conditions  
Effect on accuracy (% of span): Temperature: 0.05% per °C; RH: <1%; Line Voltage: <0.1%  
per volt; Power Loss: memory retention by EEPROM (no batteries).  
2.1.7 Outputs  
Standard alarm/control relays  
Two SPDT Form C general purpose relays rated 0.6 amps at 120/240 Vac, 0.6 amps at 110 Vdc,  
2 amps at 30 Vdc. Maximum switching power for ac, 125 VA resistive; for dc, 60W resistive.  
Optional alarm/control relays  
9782C- _ _ - R2:  
9782C- _ _ - R0:  
Two additional SPDT Form C, rated at 3 amps at 120 Vac  
or 28 Vdc, no analog outputs  
Two hermetically sealed Form C, rated at 3amps at 120 Vac  
or 28 Vdc, plus one analog output 0 to 1 Vdc or 0 to 10 Vdc  
or isolated 4 to 20 mA.  
Optional analog outputs  
One, two or three 4 to 20 mA outputs are available via an optional plug-in circuit card and  
program module. Isolated from ground, inputs and one another.  
Maximum isolation voltage: 240 Vac (will withstand 1500V hi-pot test).  
Output adjustment: to within 0.01 mA.  
Repeatability: ±0.1% of span. 600 ohm maximum load resistance. Proportional to selected  
output  
range.  
9782C- _ _ - 00:  
9782C- _ _ -VC:  
None  
One analog output: 0 to1V or 0 to10 Vdc or isolated 4 to20  
mA dc (600 ohms maximum load)  
Two analog outputs: both isolated 4 to 20 mA dc.  
Three analog outputs: all 4 to 20 mA dc  
9782C- _ _ - C2:  
9782C- _ _ - C3:  
One analog output also available in combination with optional relays (see above).  
2-4  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Model Number  
2.1.8 Temperature Compensation  
Availability  
All models can provide high resolution (0.05º C, typical) temperature compensation by  
microprocessor calculation over 0 to 140º C for high purity water and concentrations as well as  
routine water measurements.  
standard range models: Temperature compensation algorithms are for NaCl,  
acid/cation/ammonia, or morpholine.  
wide range models: Temperature compensation algorithms are for NaCl, HCl,  
NaOH, or H2SO4.  
Displayed process value can be toggled between compensated and uncompensated while the  
Analyzer/Controller is online.  
2.1.9 AutoClean  
Cell rinse schedule settings  
Timer Cycle: Adjsutable; up to 28 days with day, hour, minute resolution.  
Rinse Time: 1-1999 seconds.  
Resume Process Time Delay: 1-1999 seconds.  
2.1.10 Control  
Settings  
On/off period: 1 to 1000 seconds  
On/off percent “on” time: 0 to 100%, 1% resolution.  
Setpoint and proportional band limit ranges:conductivity: 0.055 to 19,999 µS/cm  
resistivity: 0 to 19.99 Mohms/cm  
temperature: 10 to 140 °C  
DAT cycle period: 1 to 200 seconds.  
PFT maximum frequency: 1 to 200 pulses/minute.  
PFT pulse width: 50 ms, compatible with electronic pulse-type metering pumps.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
2-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
2.1.11 Computed Values  
Available selections  
Available Computed Values  
Equation  
Description  
Range  
0.001 to 19,990  
Cell1  
Cell2  
ratio between the values  
measured by the cells  
Cell1Cell2  
difference between the values  
measured by the cells  
-19,990 to 19,990  
0 to 100%  
Cell1  
x100  
percent passage  
Cell2  
Cell1  
percent rejection  
0 to 100%  
(1−  
)x100  
Cell2  
parts per million carbon  
dioxide (9782C-S0 only)  
0 to 19.99 ppm  
2-6  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Model Number  
2.2 Model Number Breakdown  
Introduction  
The model number breakdown is presented in the tables that follow.  
The basic model number consists of a key number. Appended to this key number are characters  
that identify the features in various categories. The meaning of the characters in each category is  
presented in a table identified below. Note that because Tables III and IV do not affect I/O and  
configuration parameter selection, the characters in these tables are displayed as “don’t care”  
conditions (X) in product information viewed using the maintenance menu.  
Instructions  
9782 Analyzer Offers:  
Standard - Backlit Display  
Standard - 120/240 Volt Power Requirements  
Standard - All models CE Compliant  
Standard - All Conductivity models are 2 cell inputs  
Standard - Built-in software for ON/OFF, Duration Adjusting and Pulse  
Frequency Type Discrete Control with Standard and/or Optional relays,  
Menu Driven Prompts for Display / Programming / Maintenance,  
Text messages for Alarms/Diagnostics  
Standard - FM Haz. Loc. Approval, Class I, DIV 2 (relays rated for 120 volt max)  
Optional - Built-in software for proportional only Current Adjusting type control  
with the selection of a 4-20 mA output(s)  
9782S - Specific Ion Analyzer Offers:  
Standard - All of the above  
Standard - All Specific Ion models include one 4-20mA output and 4 relays  
(No other options from Table II)  
Select the desired key number. The arrow to the right marks the selection available.  
Make one selection from Tables using the column below the proper arrow.  
A dot ( ) denotes unrestricted availability.  
Key Number  
I
II  
III  
IV  
_ _ _ _ _  
-
_ _  
-
_ _  
-
_ _ _ _ _ _  
-
_ _  
KEY NUMBER  
Selection  
Availability  
Description  
pH/ORP  
Conductivity  
Specific Ion  
9782P  
9782C  
9782S  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
2-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
TABLE I - INPUTS  
pH/ORP Electrodes  
pH from External Pre-amp  
01  
02  
03  
04  
Direct Glass/Antimony pH/ORP Electrode/Internal Pre-amp  
Direct DURAFET pH electrode, Internal Pre-amp  
Direct Glass from HPW7000, Hi-pHurity Water Assembly,  
Internal Preamp  
(Note 2)  
Conductivity Cells  
Range  
Standard  
Wide  
S0  
W0  
(Note 3)  
(Note 4)  
Specific Ion  
Range - 0.001 to 1990  
0M  
TABLE II - AVAILABLE OUTPUTS  
No Outputs or Additional Relays  
0-1V & 0-10 V and 4-20 mA  
Two 4-20 mA  
Three 4-20 mA  
Two Additional Relays - Hermetically Sealed  
and 0-1V, 0-10V, 4-20 mA output  
00  
VC  
C2  
C3  
R0  
9782 _  
Availability  
TABLE III - OPTIONS  
Selection  
P
C
S
E _ _ _ _ _  
User’s Manual  
Pipe Mounting Kit  
English  
None  
_ 0 _ _ _ _  
_ 1 _ _ _ _  
Pipe Mounting Kit  
None  
Linen  
Tagging  
Power  
_ _ 0 _ _ _  
_ _ L _ _ _  
_ _ S _ _ _  
(Note 1)  
(Note 1)  
Stainless Steel  
Factory Set 120 Volt Operation  
_ _ _ 0 _ _  
_ _ _ 1 _ _  
Factory Set 240 Volt Operation  
(Relay contacts rated for 120 volt maximum)  
_ _ _ _ 0 _  
_ _ _ _ 1 _  
Certificate of  
Calibration  
No  
Yes  
TABLE IV  
Factory Use Only  
BD  
2-8  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications and Model Number  
Note 1:  
Customer must supply tagging information; Up to 3 lines allowed.  
(22 characters for each line)  
Note 2:  
Note 3:  
This option for use with HPW7000 assembly only.  
Standard range should be selected if measuring the following  
conductivity, resistivity or TDS ranges:  
STANDARD RANGE  
Conductivity  
0-1.999  
TDS  
Cell Constant  
0-1999 ppb  
0-19,990 ppb  
0-199.9 ppm  
0-1999 ppm  
0-10,000 ppm  
0.01  
0-19.99  
0.1  
1.0  
0-199.9  
0-1999  
10  
0-19,990  
Resistivity  
0-19.99M  
0-1.999M  
0-199.9k  
0-19.99k  
0-1.999k  
25 or 50  
Cell Constant  
0.01  
0.1  
1.0  
10  
25 or 50  
Note 4:  
Wide range should be selected if measuring the following  
conductivity, TDS or weight % concentration ranges:  
WIDE RANGE  
Conductivity  
100-1999  
TDS  
Cell Constant  
0-1999 ppm  
0-19,990 ppm  
0-199.9 ppt  
0.1  
1000-19,990  
10-199.9 mS  
10-500 mS  
1.0  
10  
25  
50  
20-1000 mS  
Weight % Concentration  
Cell Constant  
50  
0-18.00% Hydrochloric Acid  
0-19.99% Sulfuric Acid  
50  
0-19.99% Sodium Hydroxide  
0-19.99% Sodium Chloride  
25 or 50  
25 or 50  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
2-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
2-10  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Unpacking, Preparation, and Mounting  
3. Unpacking, Preparation, and Mounting  
3.1 Overview  
Introduction  
This section contains instructions for unpacking, preparing, and mounting the  
Analyzer/Controller. Instructions for wiring are provided in Sections 4 and 6. Software  
configuration is described in Section 5.  
What’s in this section?  
The topics in this section are listed below.  
Topic  
See Page  
3-1  
3.1 Overview  
3.2 Unpacking and Preparing  
3.3 Mounting  
3-2  
3-3  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
3.2 Unpacking and Preparing  
Procedure  
Table 3-1 contains procedure for unpacking and preparing the 9782.  
Table 3-1 Procedure for Unpacking and Preparing the 9782  
Step  
Action  
ATTENTION  
For prolonged storage or for shipment, the instrument should be kept in its shipping  
container. Do not remove shipping clamps or covers. Store in a suitable environment only  
(see specifications in Section 2).  
Carefully remove the instrument from the shipping container.  
1
2
Compare the contents of the shipping container with the packing list.  
Notify the carrier and Honeywell immediately if there is equipment damage or shortage.  
Do not return goods without contacting Honeywell in advance.  
Remove any shipping ties or packing material. Follow the instructions on any attached tags,  
and then remove such tags.  
3
4
All 9782 Analyzer/Controllers are calibrated and tested at the factory prior to shipment.  
Examine the model number on the nameplate (Figure 3-1) to verify that the instrument has  
the correct optional features. (See Section 2 for model number breakdown.)  
Select an installation location that meets the specifications in Section 2. The 9782 can be  
panel-, wall- or pipe-mounted (see 3.4).  
5
ATTENTION  
Pipe mounting is not recommended if the pipe is subject to severe vibration. Excessive vibration  
may affect system performance.  
If extremely hot or cold objects are near the installation location, provide radiant heat  
shielding for the instrument.  
6
3-2  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Unpacking, Preparation, and Mounting  
Honeywell  
York, PA U.S.A.  
ATTENTION - UNIT SET FOR 240 VAC  
9782C-S0-VC-E0010-00  
S/N: 9751Y712345670001  
USA WARRANTY ASSISTANCE - 1 - 800-423-9883  
120/240 VAC 50/60 Hz 15VA  
Contact Rating 1 & 2: 0.6 Amps at 120/240 VAC,  
0.6 Amps at 110 VDC, and 2 Amps at 30 VDC.  
If installed 3 & 4: 3 Amps at 120 VAC, 28 VDC.  
CE  
CASE: NEMA 4X  
Directions: 70-82-25-74  
Figure 3-1 Sample Nameplate  
3.3 Mounting  
Introduction  
The 9782 Analyzer/Controller can be panel-, wall- or pipe-mounted.  
Panel-mounting  
Figures 3-2-1 and 3-2-2 (non-backlit) and 3-3-1 and 3-3-2 (backlit) show panel cutout  
dimensions. They also illustrate how the instrument is attached to a panel by clamping the edges  
of the cutout between the case flange and the supplied U-bracket which is fastened to the rear of  
the case.  
The panel may be up to 3/ 8 in. thick. Cutouts for adjacent 9782 Analyzer/Controllers may be no  
closer than .687 in. horizontally and .937 in. vertically, resulting in a nominal edge-to-edge bezel  
separation of .078 in.  
Note that if the panel already has a cutout for a 7082 instrument, the same cutout can be used for  
the 9782. If the panel has a cutout for a 7070, 7075, 7076, 7077, or 7078 instrument, a Reducer  
Panel Kit is available for mounting the 9782. The part number is in Section 11.  
Wall- and pipe-mounting  
A bracket is available as an option for wall or pipe mounting the unit as shown in Figures 3-2-1  
and 3-2-2 (non-backlit) and 3-3-1 and 3-3-2 (backlit).  
The mounting bracket part number is in Section 11.  
ATTENTION  
Pipe mounting is not recommended if the pipe is subject to severe vibration. Excessive vibration  
may affect system performance.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
3-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
7.00 ± .040  
177.80 ± 1.02  
Case Outline  
6.12 ± .030  
155.57 ± 0.76  
Square  
(2) ¼-20 Tapped Holes  
in rear of case for  
mounting bracket  
6.00  
152.40  
C
L
4.81  
122.17  
C
L
2.41  
61.21  
.87  
22.22  
Dia holes (4)  
for lead wires,  
accomodates  
by customer  
1.37  
34.92  
Terminal Boards  
.50  
12.70  
2.75  
69.85  
conduit  
+.047  
- 0  
+1.19  
- 0  
5.516  
140.11  
inches  
millimeters  
LEGEND:  
min space  
between  
horizontal  
cutouts  
.687  
17.45  
2.76  
70.05  
+.047  
5.516  
140.11  
- 0  
+1.19  
- 0  
.125  
3.18 R max  
2.76  
70.05  
permissible  
.937  
23.81  
Panel Cutout  
min space between  
vertical cutouts  
7.43 ± .070  
188.72 ± 1.78  
7.68 ± .060  
195.07 ± 1.52  
3.06  
77.80  
1.00  
IPS Pipe  
(by customer)  
25.40  
Note 1: Do not exceed 80 lb-in torque  
when tightening fastners.  
Horizontal Rear Pipe Mounting  
Figure 3-2-1 Mounting and Dimensions – Non-Backlit Display Option  
3-4  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Unpacking, Preparation, and Mounting  
inches  
millimeters  
LEGEND:  
Vertical Rear Pipe Mounting  
3.06  
77.80  
7.43 ±.070  
188.72 ±1.78  
7.68 ±.060  
195.07 ±1.52  
1.00  
25.40  
IPS Pipe  
(by customer)  
Note 1: Do not exceed 80 lb-in torque  
when tightening fastners.  
Top View  
Panel Mounting  
into existing 7075, 7076, 7077, and 7078 series cutouts  
7.43 ±.070  
1.50  
188.72 ±1.78  
38.10  
C
L
3.00  
76.20  
C
L
6.43 ±0.70  
163.32 ±1.78  
1.06 ±.025  
26.92 ±0.64  
6.27 ±.040  
159.26 ±1.02  
.25 max.  
6.35  
Customer Panel  
Four holes in bracket or  
1/4 dia. mounting bolts  
(bolts by customer)  
R.H. Side View  
Wall Mounting  
Panel Mounting  
6.63  
168.40  
C
L
6.43 ±0.70  
163.32 ±1.78  
C
L
3.31  
84.07  
1.00 ±.025  
25.40 ±0.64  
.12 min.  
3.04  
6.27 ±.040  
159.26 ±1.02  
.37 max.  
9.40  
7.68 ±.060  
195.07 ±1.52  
Customer Panel  
Figure 3-2-2 Mounting and Dimensions – Non-Backlit Display Option  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
3-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
7.21 ± .040  
183.13 ± 1.02  
Case Outline  
6.12 ± .030  
155.57 ± 0.76  
Square  
(2) ¼-20 Tapped Holes  
in rear of case for  
mounting bracket  
6.00  
152.40  
C
L
4.81  
122.17  
C
L
2.41  
61.21  
.87  
22.22  
Dia holes (4)  
for lead wires,  
accomodates  
by customer  
1.37  
34.92  
Terminal Boards  
.50  
12.70  
2.75  
69.85  
conduit  
+.047  
- 0  
+1.19  
- 0  
5.516  
140.11  
inches  
millimeters  
LEGEND:  
min space  
between  
horizontal  
cutouts  
.687  
17.45  
2.76  
70.05  
+.047  
5.516  
140.11  
- 0  
+1.19  
- 0  
.125  
3.18 R max  
2.76  
70.05  
permissible  
.937  
23.81  
Panel Cutout  
min space between  
vertical cutouts  
7.64 ± .070  
194.06 ± 1.78  
7.68 ± .060  
195.07 ± 1.52  
3.06  
77.80  
1.00  
IPS Pipe  
(by customer)  
25.40  
Note 1: Do not exceed 80 lb-in torque  
when tightening fastners.  
Horizontal Rear Pipe Mounting  
Figure 3-3-1 Mounting and Dimensions – Backlit Display Option  
3-6  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Unpacking, Preparation, and Mounting  
inches  
millimeters  
Vertical Rear Pipe Mounting  
LEGEND:  
3.06  
77.80  
7.64 ±.070  
194.06 ±1.78  
7.68 ±.060  
195.07 ±1.52  
1.00  
IPS Pipe  
(by customer)  
25.40  
Note 1: Do not exceed 80 lb-in torque  
when tightening fastners.  
Top View  
Panel Mounting  
into existing 7075, 7076, 7077, and 7078 series cutouts  
7.64 ±.070  
1.50  
194.06 ±1.78  
38.10  
C
L
3.00  
76.20  
C
L
6.43 ±0.70  
163.32 ±1.78  
1.27 ±.025  
32.26 ±0.64  
6.27 ±.040  
159.26 ±1.02  
.25 max.  
6.35  
Customer Panel  
Four holes in bracket or  
1/4 dia. mounting bolts  
(bolts by customer)  
R.H. Side View  
Wall Mounting  
Panel Mounting  
6.63  
168.40  
C
L
6.43 ±0.70  
163.32 ±1.78  
C
L
3.31  
84.07  
1.21 ±.025  
30.73 ±0.64  
.12 min.  
3.04  
6.27 ±.040  
159.26 ±1.02  
.37 max.  
9.40  
7.68 ±.060  
195.07 ±1.52  
Customer Panel  
Figure 3-3-2 Mounting and Dimensions – Backlit Display Option  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
3-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
3-8  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Power Wiring  
4. Power Wiring  
4.1 Overview  
Introduction  
This section contains instructions for installing ac power wiring for the Analyzer/Controller, in  
preparation for configuring the software and performing I/O setup as described in Section 5.  
We recommend that you wait to install input and output wiring (see Section 6) until after I/O  
setup. During I/O setup the software will determine for you which relay to use for each feature.  
What’s in this section?  
The topics in this section are listed below.  
Topic  
See Page  
4-1  
4.1 Overview  
4.2 General Wiring Practices for Power Wiring  
4.3 Power Wiring Considerations  
4.4 Installing Power Wiring  
4-2  
4-3  
4-4  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
4-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
4.2 General Wiring Practices for Power Wiring  
WARNING  
Wiring should be performed by qualified personnel only.  
Safety precautions  
WARNING  
A disconnect switch must be installed to break all current carrying conductors. Turn off power  
before working on conductors. Failure to observe this precaution may result in serious personal  
injury.  
WARNING  
An external disconnect switch is required for any hazardous voltage connections to the relay outputs.  
Avoid damage to components  
ATTENTION  
This equipment contains devices that can be damaged by electrostatic discharge (ESD). As  
solid state technology advances and as solid state devices get smaller, they become  
increasingly sensitive to ESD. The damage incurred may not cause the device to fail  
completely, but may cause early failure. Therefore, it is imperative that assemblies containing  
static sensitive devices be carried in conductive plastic bags. When adjusting or performing any  
work on such assemblies, grounded work stations and wrist straps must be used. If soldering  
irons are used, they must also be grounded.  
A grounded work station is any conductive or metallic surface connected to an earth ground,  
such as a water pipe, with a 1/2 to 1 megohm resistor in series with the ground connection. The  
purpose of the resistor is to current limit an electrostatic discharge and to prevent any shock  
hazard to the operator. The steps indicated above must be followed to prevent damage and/or  
degradation, which may be induced by ESD, to static sensitive devices.  
Wiring for immunity compliance  
In applications where either the power, input or output wiring are subject to electromagnetic  
disturbances, shielding techniques will be required. Grounded metal conduit with conductive  
conduit fittings is recommended.  
Connect the ac mains through a fused disconnect switch. To ensure that the unit meets the  
requirements of the EMC directive 89/336/EEC, a power line filter kit is included with each unit.  
Wind three turns of each wire (14 AWG maximum) through the filter core for the two power and  
protective earth leads as illustrated in Figure 4-1. The wound filter cores shall be located  
external to the 9782 case within 25 cm.  
Conform to code  
Instrument wiring should conform to regulations of the National Electrical Code.  
4-2  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Power Wiring  
4.3 Power Wiring Considerations  
Recommended wire size  
Observe all national and local electrical codes when making power connections. Unless locally  
applicable codes dictate otherwise, use 14 gauge (2.081 mm2) wire for ac power.  
Power supply voltage and frequency within specs  
The power supply voltage and frequency must be within the limits stated in the specifications in  
Section 2.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
4-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
4.4 Installing Power Wiring  
Procedure  
Follow the procedure in Table 4-1 to install AC power wiring.  
WARNING  
Turn power off at mains before installing AC power wiring.  
Table 4-1 Procedure for Installing AC Power Wiring  
Action  
Step  
1
Check the tag on the outside of the case to be sure that the voltage rating of the unit  
matches the supply voltage at your site.  
ATTENTION  
The unit may be damaged if you apply power with the wrong voltage.  
Open the case:  
2
Grasp the bottom center portion of the front bezel and pull it downward and toward you  
slightly to disengage the bottom of the bezel from the edge of the case.  
Lift the bezel gently to disengage it from the top edge of the case.  
Swing the bezel to the left. (The bezel and display assembly is mounted on pivot arms.)  
The safety cover with its warning label will be visible.  
Remove the safety cover by removing the single screw holding it in place.  
3
4
Install a fused disconnect switch in the power line which will be connected to the  
Analyzer/Controller.  
If a 230/240 Vac line is to be connected, use a 0.125 amp fuse.  
If a 110/120 Vac line is to be connected, use a 0.25 amp fuse.  
To ensure that the unit meets the requirements of EMC directive 89/336/EEC, install the  
power line filter kit supplied with the Analyzer/Controller. Wind three turns of each wire (14  
AWG maximum) through the filter core for the two power and protective earth leads as  
illustrated in Figure 4-1. The wound filter cores shall be located external to the 9782 case  
within 25 cm.  
5
6
Connect the power wiring to terminals L1 and L2/N as shown in Figure 4-1. Connect the hot  
side of the line to terminal L1 and the grounded side of the line to terminal L2/N.  
If the L2/N side of the line cannot be grounded, add an isolating transformer electrically  
between the power supply and the instrument and ground the L2/N terminal.  
WARNING  
The ground terminal must be connected to a reliable earth ground. If metal conduit is used, connect  
a bonding wire between conduits. Do not rely upon the conductive coating of the instrument case to  
provide this connection. Failure to observe this precaution may result in serious personal injury.  
Replace the safety cover and secure it with the screw.  
7
8
Close the case and power up the unit. Do not apply power until the case is closed.  
4-4  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Power Wiring  
Power Terminations  
L1  
L2/N  
PE  
PE = Protective  
Earth Ground  
Figure 4-1 AC Power Terminals  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
4-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
4-6  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
5. I/O Setup and System Configuration  
5.1 Overview  
Introduction  
This section provides instructions for performing I/O setup and for configuring the  
Analyzer/Controller’s software to implement your control strategy.  
What’s in this section?  
The topics in this section are listed below.  
Topic  
See Page  
5-1  
5.1 Overview  
5.2 I/O Setup and Configuration Tasks  
5.3 Performing I/O Setup  
5-2  
5-4  
5.4 Configuring Current Adjusting Type Control and/or Retransmission of Process  
Variables  
5-10  
5.5 Configuring On/Off, Duration Adjusting Type (DAT) or Pulse Frequency Type  
(PFT) Control  
5-12  
5.6 Configuring Alarms  
5-16  
5-17  
5-18  
5-21  
5-23  
5.7 Setting the Clock  
5.8 Configuring AutoClean  
5.9 Selecting a Computed Value To Be Displayed  
5.10 Summary of Menu Choices and System Defaults  
Powering up the Analyzer/Controller  
Every time the Analyzer/Controller is powered up the unit performs a “warm reset”. During a  
warm reset, alarms are not cleared, and outputs are maintained at their last value prior to the loss  
of power or initiation of the warm reset. A warm reset has no effect on configuration or  
calibration values.  
In addition to this warm reset, a “cold reset” function is available to return ALL configuration  
values to the factory settings shown in Table 5-10. During a cold reset some I/O setup values  
will also be changed. In addition, the cal factor will be set to match the default cell constant, and  
the calibration trim or TDS ( total dissolved solids) conversion factor will be removed,  
eliminating any changes made using the calibration menu (see Section 8). All alarms will be  
cleared. Analog outputs will go to zero momentarily, then go to the factory default values. A  
cold reset can be initiated using a maintenance menu function as described in Section 10. It will  
also be initiated during I/O setup if the measurement type, relay activation, display resolution,  
cell constant, or an analog or discrete I/O assignment is changed.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
5.2 I/O Setup and Configuration Tasks  
Ease of configuration  
Configuring the 9782 software is fast and easy. A menu is provided for every configuration task.  
You will be permitted to configure only those parameters relevant to your application and  
supported by the Analyzer/Controller model you purchased. In fact, configuration screens will  
contain only prompts and menu choices that apply to your application.  
Accessing the I/O setup and configuration menus  
Use the MENU key to access the main menu which includes CONFIGURATION and I/O  
SETUP choices.  
If the security feature has been enabled as described in Section 10, the password must be entered  
in response to the prompt before you can proceed. (Out-of-the-box units do not require entry of a  
password to access the menu system.)  
Purpose of I/O setup  
I/O setup can be accessed to either choose or review I/O parameters. When you choose features  
during I/O setup, the software makes appropriate I/O assignments. I/O setup (see 5.3) must  
precede configuration. The choices made during I/O setup determine which prompts and menu  
choices are available during configuration. (Even though I/O setup is performed first, I/O  
SETUP is last on the main menu because you usually only have to access this function a few  
times in the life of your Analyzer/Controller.)  
Purpose of configuration menu items  
The configuration menu contains the following items for accessing screens used to configure the  
features described below.  
CAT/RETRANSMISSION: If the hardware supports analog outputs, configure parameters  
related to use of these outputs as described in 5.4.  
ALARMS: If one or more relays are available for annunciation of alarms, configure each  
alarm input, setpoint, deadband, and delay as described in 5.6. (Alarms are assigned to  
relays with I/O SETUP.)  
DISCRETE CONTROL: One or more relays can be used to achieve control of process  
variables using On/Off, Duration Adjusting Type (DAT) or Pulse Frequency Type (PFT)  
control strategies. Configure the related parameter values as described in 5.5.  
ADVANCED FEATURES: This menu item provides access to configuration screens for  
the following features:  
CLOCK SETUP: A display showing the current time can be accessed when the unit  
is online. The 24-hour clock must be set before you can use the AutoClean feature  
described below. Set the clock as described in 5.7.  
5-2  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
AUTOCLEAN: AutoClean is a feature that lets the Analyzer/Controller periodically  
rinse the cells automatically. Configure control of AutoClean as described in 5.8.  
COMPUTED VALUE: The Analyzer/Controller can compute and display a value,  
such as the ratio between Cell 1 and Cell 2 conductivity values or the parts per million  
of carbon dioxide. Select the value to be displayed as described in 5.9. (A different  
computed value can be sent to an analog output as described in 5.3.) If a computed  
value is used, it is significant which cell is considered “Cell 1” and which is “Cell 2”.  
This must be taken into account when wiring the inputs. See 6.3 for details.  
Illustrations  
The procedures in this section contains illustrations of display screens. These pictures are  
intended to show the types of information displayed on each screen. However, the screens you  
will actually see on your Analyzer/Controller may be different from the pictures in the manual.  
The displays will be customized for the hardware and application.  
System defaults  
Table 5-10 contains the menu choices available (or ranges for numeric values) and system  
defaults for I/O assignments and configuration parameters, as well as for values entered during  
maintenance functions described in Section 10.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
5.3 Performing I/O Setup  
Purpose  
Before you can configure your Analyzer/Controller to implement your alarm and control strategy  
you must perform I/O setup. In I/O setup you choose the features to be used, based on the  
hardware capability of the 9782 model being configured. Using your selections, the software  
will display the relay and analog output assignments appropriate for your application. The  
choices made during I/O setup determine which prompts and menu choices are available during  
configuration. For a discussion of the features available, and of the interactions affecting relay  
assignments, see Section 1.  
I/O setup is also used to specify relay action. The relay action used for alarm and/or control  
action for all relays in the Analyzer/Controller can be specified as energized or de-energized on  
alarm (or when discrete control point is ON).  
In addition, I/O setup is used to specify display resolution and unit of measure (see below).  
Display resolution and unit of measure  
For standard range models (9782C-S0), the resolution of the display of conductivity or resistivity  
can be selected. (No other displays are affected.)  
When set to LOW, the decimal point is fixed. It remains in the same place throughout the  
display range.  
When set to HIGH, the decimal point will automatically be shifted one place to the left to  
provide greater resolution when the displayed value falls below 10% of the high display  
range, and conversely, be shifted back when the value increases above 10%.  
This selection of low or high display range has no effect on the fixed or auto output range status.  
For conductivity measurement the unit of measure can also be specified for standard range  
Analyzer/Controllers. No unit of measure is displayed for wide range models. The  
Analyzer/Controller automatically uses the conductivity unit of measure appropriate for the cell  
constant (see Wide Display Ranges table in Sub-Section 2.1.5.)  
ATTENTION  
If you use I/O setup to view parameters, making no changes, a cold reset will not occur.  
However, if you change any parameter other than the type of solution for which temperature  
compensation should be applied, you will cause the Analyzer/Controller to perform a cold reset  
(see page 5-1). This cold reset is necessary to ensure that the integrity of the unit remains intact  
when critical parameters have been changed.  
ATTENTION  
AutoClean and range status indication cannot be selected for use with two cells simultaneously.  
Your choices are:  
For Cell 1 only, use AutoClean (Relay 1) and/or range status indication (Relay 2).  
For both Cell 1 and Cell 2, use AutoClean (Relays 1 and 3), with no range status indication,  
nor any discrete control.  
For both Cell 1 and Cell 2, use range status indication (Relays 2 and 4), with no AutoClean,  
nor any discrete control.  
5-4  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
ATTENTION  
If you add an option card to the 9782, it is very important that you change the model number  
stored in memory (see Section 10).  
During I/O setup the software will display prompts only for those features the hardware can  
support. The model number of the unit was stored in the 9782 memory at the factory. Each  
combination of features has a unique model number (see Section 2). Therefore, the processor  
can use this information to determine which prompts and selections are appropriate.  
Entering the new number when you modify the hardware will let the processor know about the  
new hardware capabilities. As a result, you will be able to configure the unit to take advantage of  
features requiring that hardware.  
Procedure  
The procedure for I/O setup configuration is shown in Table 5-1.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Table 5-1 Procedure for Performing I/O Setup  
Step  
1
Screen  
Action  
MENU  
Go to I/O SETUP.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
RELAY SETTINGS  
2
The first screen shows the present use of available relays.  
This information is read-only.  
RLY1 RLY2 RLY3 RLY4  
ACLN RNGS ALM3 ALM4  
In this example the 9782 has four relays. Relay 1 is used  
for AutoClean. Relay 2 is used for indication of automatic  
output range switching. Relay 3 is used to trigger an alarm  
annunciator for Alarm 3. Relay 4 is used for Alarm 4  
annunciation.  
DIAGNOSTICS  
| PREV | NEXT  
Go to the next screen.  
ANALOG OUTPUTS  
3
4
The next screen shows the present assignments for the  
optional current or voltage outputs, if available. This  
information is read-only.1  
OUTPUT1  
OUTPUT2  
OUTPUT3  
C1  
C2  
RATO  
DIAGNOSTICS  
| PREV | NEXT  
Go to the next screen.  
ATTENTION  
RECONFIG OF ANY I/O  
SETUP PARM WILL  
RESET CONFIG PARMS  
TO FACTORY DEFAULTS  
MENU TO EXIT  
I/O setup is normally performed only once, before field  
connections are made to the Analyzer/Controller. You can  
also use this menu to view parameters. If you do not  
change anything, you will not cause a cold reset of the  
unit. You can exit now using the MENU key. If you go  
past this “Attention” screen, you must go through the entire  
sequence of I/O setup screens before returning to the  
main menu.  
DIAGNOSTICS  
| PREV | NEXT  
Go to the next screen.  
ADVANCED FEATURES  
5
Specify whether AutoClean should be enabled. If YES, a  
relay will be dedicated to this use.  
AUTOCLEAN  
RANGE STATUS  
YES  
NO  
Specify whether you want to use a relay to indicate when  
the range shifts for the analog output based on Cell 1’s  
input. (Output scaling is specified in Step 13.) If YES, a  
relay will be assigned to indicate when the Cell 1 range  
switch occurs.  
DIAGNOSTICS  
|
| NEXT  
Go to the next screen.  
1 If the system uses a DE option card, no analog outputs are supported by the hardware. The screen will show “DE  
ENABLED”, instead of displaying output assignments.  
5-6  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
Action  
Step  
6
Screen  
DISCRETE CONTROL  
If any relays are still available, the screen will display a  
query about discrete control.  
SELECT  
YES  
DIAGNOSTICS  
| PREV | NEXT  
DISCRETE CONTROL TYPE  
7
8
If discrete control will be used, select the relay(s) to be  
used and the type of control for each.  
RELAY 3  
RELAY 4  
N/A  
DAT  
Go to the next screen.  
DIAGNOSTICS  
| PREV | NEXT  
RELAY ASSIGNMENTS  
If any relays are still available, then indicate which feature  
will be associated with each available relay. In our  
example, only Relay 3 can be assigned using this screen.  
Relays 1, 2 and 4 have already been dedicated to other  
functions.  
RLY1 RLY2 RLY3 RLY4  
ACLN RNGS AA DAT  
DIAGNOSTICS  
| PREV | NEXT  
However, by selecting AA, all alarms can be directed to a  
single relay.  
RELAY ACTIVATION  
STATE ON ALARM  
9
The display will indicate whether relay activation in case of  
alarm (or control output “ON” state) is currently ENERGIZE  
or DE-ENERGIZE.  
SELECT DE-ENERGIZE  
Use this display to specify the desired relay action.  
DIAGNOSTICS  
| PREV | NEXT  
MEASUREMENT TYPE  
10  
Specify the type of measurement to be performed by the  
Analyzer/Controller.  
SELECT  
COND/TDS  
If you specify any measurement type other than the default  
(conductivity/total dissolved solids), the unit will  
immediately perform a cold reset, returning all  
configuration and calibration trim (or TDS factor) values to  
the factory defaults.  
DIAGNOSTICS  
| PREV | NEXT  
A cold reset will also return some I/O setup values to  
factory settings. If you change the measurement type,  
be sure to cycle through all the other I/O setup  
screens and check that the settings are appropriate  
for your application.  
Go to the next screen.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Step  
11  
Screen  
Action  
DISPLAY RESOLUTION  
RESOLUTION  
UOM  
Use this screen to specify the display resolution, LOW  
(fixed) or HIGH (automatically adjusted).  
LOW  
uS  
Also use it to specify the unit of measure for conductivity  
measurement with standard range models.  
DIAGNOSTICS  
| PREV | NEXT  
If the instrument is a standard range model and the cell  
constant is 0.01, 0.1, or 1.0, then the choices are  
microSiemens per centimeter (µS/cm) displayed as uS or  
microSiemens per meter (µS/m) displayed as uSm.  
If the instrument is a standard range model and the cell  
constant is 10 or greater, then the unit of measure is  
always microSiemens per centimeter (µS/cm) displayed as  
uS.  
SELECT CELL CONSTANT  
C1 0.1  
C2 N/A  
12  
Specify the cell constant for each cell. Note that this is the  
cell constant for the type of cell. The number can be found  
on the label on the potted connector housing of the cell.  
The cell constant is indicated by the three characters  
following the four-digit series number. For example, a cell  
labeled 4973-001 has a 0.01 cell constant. When in  
doubt, refer to the manual supplied with the cell. 2  
DIAGNOSTICS  
| PREV | NEXT  
Changing the cell constant causes the unit to perform a  
cold reset, returning all configuration and calibration trim  
(or TDS factor) values to the factory defaults.  
A cold reset will also return some I/O setup values to  
factory settings. If you change the measurement type,  
be sure to cycle through all the other I/O setup  
screens and check that the settings are appropriate  
for your application.  
If only one cell will be used, enter its constant for Cell 1  
and select N/A for Cell 2. Step 5 in Table 6-2 outlines the  
necessary wiring procedure when using Cell 1 and not Cell  
2.  
Go to the next screen.  
2
Do not confuse this cell constant, universal for the type of cell, with the cell calibration factor that applies only to  
each individual cell. The cell calibration factor is found on the plastic tag hanging from the cell lead wires. It is  
entered using the calibration menu as described in Section 8.  
5-8  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
Action  
Step  
13  
Screen  
SOLUTION TEMP COMP  
Select the type of solution for which temperature  
C1  
C2  
ACID/NH3/ETA  
NACL  
compensation should be applied.  
DIAGNOSTICS  
| PREV | NEXT  
ANALOG OUTPUTS  
14  
15  
If the hardware supports analog outputs, specify for each  
which process variable or computed value will be  
associated with the output.  
OUTPUT1  
OUTPUT2  
OUTPUT3  
C1  
C2  
RATO  
3
DIAGNOSTICS  
| PREV | NEXT  
OUTPUT SCALING  
OUTPUT1 LINFIXEDRNG  
If analog outputs will be used, specify the scaling for each  
output. 4  
OUTPUT2  
OUTPUT3  
LINAUTORNG  
BI-LINEAR  
CAT (Current Adjusting Type) control uses Fixed  
Linear Range only. Retransmitted process values can  
be scaled. A retransmitted computed value always  
uses Fixed Linear Range.  
DIAGNOSTICS  
| PREV | NEXT  
16  
At this point, you have completed the I/O setup menu. If  
you have changed at least one parameter (other than  
solution temperature compensation) the unit will perform a  
cold reset (see page 5-1).  
I/O SETUP COMPLETE  
PRESS  
ENTER TO CONTINUE  
DIAGNOSTICS  
| PREV |  
UPDATING EEPROM  
When you press ENTER to continue, the message  
“UPDATING EEPROM” will provide feedback, letting you  
know your selections have been activated.  
DIAGNOSTICS  
|
|
UPDATING EEPROM  
PRESS ENTER  
TO RETURN TO  
MAIN MENU  
When the update has been completed, you will be  
prompted to exit the menu.  
DIAGNOSTICS  
|
|
3 If the system uses a DE option card, no analog outputs are supported by the hardware. The screen will show  
“DE ENABLED”, instead of displaying output assignments.  
4
This output range scaling has no effect on the display. Display resolution is configured using the maintenance  
menu as described in Section 10.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
5.4 Configuring Current Adjusting Type (CAT) Control and/or  
Retransmission of Process Variables  
Purpose  
If the hardware supports analog outputs, during I/O setup you had an opportunity to specify  
which process variable or computed value will be associated with each output. An analog output  
can be used for scaled retransmission of the value, or for Current Adjusting Type control by  
means of an output signal that is directly proportional to the input signal or computed value.  
Before using analog outputs, you must configure the related parameter values.  
For a discussion of tuning proportional control see Appendix A.  
Procedure  
The procedure for configuring CAT control and/or retransmission of process variables is shown  
in Table 5-2.  
Table 5-2 Procedure for Configuring CAT Control and/or Retransmission of  
Process Variables  
Step  
1
Screen  
Action  
MENU  
Go to CONFIGURATION | CAT/RETRANSMISSION.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
CONFIGURATION  
CAT/RETRANSMISSION  
ALARMS  
DISCRETE CONTROL  
ADVANCED FEATURES  
ATTENTION: Any value entered for one of these  
parameters will take effect as soon as it is entered.  
DIAGNOSTICS  
| PREV | NEXT  
CAT/RETRANSMISSION  
CAT  
2
If the Analyzer/Controller hardware only supports one  
analog output, the screen at the left will be displayed when  
CAT/RETRANSMISSION is selected from the menu.5  
RETRANSMISSION  
DIAGNOSTICS  
| PREV | NEXT  
If the hardware supports three analog outputs, the screen  
in Step 6 will be displayed when CAT/RETRANSMISSION  
is selected from the menu.  
If CAT control will be used, select it from the menu.  
5 If the system uses a DE option card, no analog outputs are supported by the hardware. The screen will show “DE  
ENABLED”, instead of displaying analog outputs.  
5-10  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
Action  
Step  
3
Screen  
CAT CONTROL  
Select an output to configure it for control.  
CAT1  
CAT2  
CAT3  
Go to the next screen.  
DIAGNOSTICS  
| PREV | NEXT  
CAT CONTROL OUTPUT 1  
SCALING LINFIXEDRNG  
4
The input type and appropriate unit of measure will already  
be filled in, based on the process variable or computed  
value assigned to this analog output during I/O setup. The  
scaling characteristic will also be filled in; CAT output is  
always linear fixed-range. (All this is read-only  
information.)  
INPUT  
SETPOINT  
PB LIMIT  
C1  
12000µS  
300µS  
DIAGNOSTICS  
| PREV | NEXT  
Use the setpoint and proportional band parameters to  
define the output span.  
RETRANSMISSION  
OUTPUT 1  
OUTPUT 2  
5
6
To configure retransmission on a unit with a single analog  
output, go to CONFIGURATION |  
CAT/RETRANSMISSION | RETRANSMISSION.  
DIAGNOSTICS  
| PREV | NEXT  
To configure retransmission on a unit with three analog  
outputs go to CONFIGURATION |  
CAT/RETRANSMISSION and select the output to be  
configured.  
OUTPUT 1  
SCALING LINFIXEDRNG  
For each output to which a process variable or computed  
value was assigned during I/O setup, the value assigned  
will be displayed, as well as the type of scaling to be  
applied, and the appropriate unit of measure. (This is  
read-only information.)  
INPUT  
0%  
100%  
C1  
0.0µS  
19.0µS  
DIAGNOSTICS  
| PREV | NEXT  
Specify the range by entering the values to be represented  
by outputs of 0% and 100%. If reverse-acting output is  
required, enter the high range value at the 0% prompt and  
the low range value at the 100% prompt.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
5.5 Configuring On/Off, Duration Adjusting Type (DAT), or  
Pulse Frequency Type (PFT) Control  
Purpose  
During I/O setup you had an opportunity to specify whether to use one or more relays to achieve  
control of a process variable with a discrete control strategy: On/Off, DAT or PFT.  
On/Off control can be implemented based on any measured or computed value.  
Cell1  
DAT and PFT control can be based on Cell 1 input, Cell 2 input, the computed ratio Cell2 , or the  
Cell1Cell2  
computed difference  
.
If you specified one or more of these discrete control types during I/O setup, you will be able to  
configure the related parameter values.  
For a discussion of tuning proportional control (applicable to PFT and DAT control) see  
Appendix A.  
Procedure  
The procedure for configuring On/Off control is shown in Table 5-3.  
The procedure for configuring PFT control is shown in Table 5-4.  
The procedure for configuring DAT control is shown in Table 5-5.  
5-12  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
Table 5-3 Procedure for Configuring On/Off Control  
Step  
1
Screen  
Action  
MENU  
Go to CONFIGURATION | DISCRETE CONTROL |  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
ON/OFF.  
DIAGNOSTICS  
| PREV | NEXT  
CONFIGURATION  
CAT/RETRANSMISSION  
ALARMS  
ATTENTION: Any value entered for one of these control-  
related parameters will take effect as soon as it is entered.  
DISCRETE CONTROL  
ADVANCED FEATURES  
DIAGNOSTICS  
| PREV | NEXT  
DISCRETE CONTROL  
ON/OFF  
PFT  
DAT  
DIAGNOSTICS  
|
|
ON/OFF CONTROL  
CYCLE PERIOD  
CYCLE TMR 1  
CYCLE TMR 2  
CYCLE TMR 3  
CYCLE TMR 4  
DIAGNOSTICS  
2
3
Use this screen to specify the length of the cycle period for  
all timers.  
500s  
To configure an individual timer associated with one of the  
relays, select it from the menu.  
|
|
CYCLE TIMER 1 RLY2  
INPUT  
The relay assigned to this On/Off control output during I/O  
setup will already be filled in. (This is read-only  
information.)  
C1  
HIGH  
ACTION  
SETPOINT  
DEADBAND  
ONTIME  
15000 µS  
250 µS  
50%  
DIAGNOSTICS  
Specify the process value or computed value to be used  
as the input.  
|
|
Specify whether relay activation should occur when the  
input’s value exceeds the setpoint (HIGH action) or falls  
below the setpoint (LOW action).  
Enter the setpoint.  
If desired, specify a deadband. (Once the output is ON, it  
will remain ON until the controlled variable differs from the  
setpoint by at least the value of the deadband.)  
Enter the percent of the cycle period that the relay should  
stay on when the setpoint is reached.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Table 5-4 Procedure for Configuring PFT Control  
Step  
1
Screen  
Action  
MENU  
Go to CONFIGURATION | DISCRETE CONTROL | PFT.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
CONFIGURATION  
CAT/RETRANSMISSION  
ALARMS  
ATTENTION: Any value entered for one of these control-  
related parameters will take effect as soon as it is entered.  
DISCRETE CONTROL  
ADVANCED FEATURES  
DIAGNOSTICS  
| PREV | NEXT  
DISCRETE CONTROL  
ON/OFF  
PFT  
DAT  
DIAGNOSTICS  
|
|
PFT CONTROL  
DB FREQ RATE 100 P/M  
2
3
Specify the maximum frequency rate (in pulses per  
minute) representing 100% output. (Be sure this rate is  
compatible with the controlled device.) This rate will be  
used for all PFT outputs in the Analyzer/Controller.  
PFT1  
PFT2  
DIAGNOSTICS  
|
|
Use this screen to select the PFT control output to be  
configured.  
PFT 1  
The relay assigned to this PFT control output during I/O  
setup will already be filled in. (This is read-only  
information.)  
OUTPUT  
INPUT  
SETPOINT  
RLY2  
C1  
15000 µS  
250 µS  
PB LIMIT  
DIAGNOSTICS  
|
Specify the process value or computed value to be used  
as the input.  
|
Enter the setpoint at which the output should be turned on.  
Enter the PB Limit. This is the value for which the output’s  
pulse output rate will be at its maximum frequency.  
5-14  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
Table 5-5 Procedure for Configuring DAT Control  
Step  
1
Screen  
Action  
MENU  
Go to CONFIGURATION | DISCRETE CONTROL | DAT.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
CONFIGURATION  
CAT/RETRANSMISSION  
ALARMS  
ATTENTION: Any value entered for one of these control-  
related parameters will take effect as soon as it is entered.  
DISCRETE CONTROL  
ADVANCED FEATURES  
DIAGNOSTICS  
| PREV | NEXT  
DISCRETE CONTROL  
ON/OFF  
PFT  
DAT  
DIAGNOSTICS  
|
|
DAT CONTROL  
CYCLE PERIOD 200 S  
DAT1  
DAT2  
2
3
Specify the length of the cycle period (in seconds) for the  
time-proportioned control.  
Use this screen to select the DAT control output to be  
configured.  
DIAGNOSTICS  
|
|
ON/OFF CYCLE TIMER 1  
RLY2  
C1  
The relay assigned to this DAT control output during I/O  
setup will already be filled in. (This is read-only  
information.)  
OUTPUT  
INPUT  
SETPOINT  
PB LIMIT  
15000 µS  
250 µS  
DIAGNOSTICS  
|
Specify the process value or computed value to be used  
as the input.  
|
Enter the setpoint at which the output should be turned on.  
Enter the PB Limit. This is the value for which the output’s  
on duration will be maximum.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
5.6 Configuring Alarms  
Purpose  
If you assigned one or more alarms to at least one relay during I/O setup (described in 5.3), then  
for each alarm select the process value or computed value to be alarmed, setpoint, deadband, and  
delay as described below.  
Procedure  
The procedure for configuring parameters for an alarm is described in Table 5-6.  
Table 5-6 Procedure for Configuring Alarms  
Step  
1
Screen  
Action  
MENU  
Go to CONFIGURATION | ALARMS.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
CONFIGURATION  
CAT/RETRANSMISSION  
ALARMS  
ATTENTION: Any value entered for one of these alarm-  
related parameters will take effect as soon as it is entered.  
DISCRETE CONTROL  
ADVANCED FEATURES  
DIAGNOSTICS  
| PREV | NEXT  
ALARMS  
DELAY TIME  
ALARM1  
2
3
Use this screen to enter a delay time to prevent brief  
process upsets from activating an alarm. The delay  
entered here will apply to all alarms.  
3 S  
ALARM2  
ALARM3  
ALARM4  
DIAGNOSTICS  
|
Select the alarm to be configured.  
|
ALARM1  
The relay to which the alarm was assigned during I/O  
setup will be displayed as read-only information.  
OUTPUT  
INPUT  
ACTION  
SP  
RLY2  
C1  
HIGH  
15000 µS  
250 µS  
Enter the measured variable or computed value to be  
alarmed.  
DEADBAND  
DIAGNOSTICS  
|
|
Indicate whether this is a high or low alarm.  
Enter the setpoint value at which the alarm will become  
active.  
Specify the deadband. Once the alarm is active, it will not  
deactivate until the alarmed variable differs from the  
setpoint by at least the value of the deadband.  
5-16  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
5.7 Setting the Clock  
Purpose  
A display showing the current time can be accessed when the unit is online. The 24-hour clock  
must be set before you can use the AutoClean feature described in Sub-Section 5.6.  
Procedure  
Follow the procedure in Table 5-7 to set the 24-hour clock.  
Table 5-7 Procedure for Setting the Clock  
Step  
1
Screen  
Action  
MENU  
Go to CONFIGURATION | ADVANCED FEATURES |  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
CLOCK SETUP.  
DIAGNOSTICS  
| PREV | NEXT  
CONFIGURATION  
CAT/RETRANSMISSION  
ALARMS  
DISCRETE CONTROL  
ADVANCED FEATURES  
DIAGNOSTICS  
| PREV | NEXT  
ADVANCED FEATURES  
CLOCK SETUP  
COMPUTED VALUE  
AUTO CLEAN  
DIAGNOSTICS  
|
|
24 HOUR CLOCK SETUP  
2
Set the clock. For example, if you will start the clock at  
1:30 PM, enter 13 hours, 30 minutes.  
CLOCK  
HOURS  
MINUTES  
OFF  
13  
30  
Start the clock and specify what should happen to the  
clock if the power fails.  
DIAGNOSTICS  
|
| NEXT  
ON W/RESET starts the clock now. Following power  
failure, the clock will be reset to 00:00 and the diagnostic  
message “RESET CLOCK” will be displayed.  
ON W/O RESET starts the clock now. Following power  
failure the clock will remain off, and the diagnostic  
message “RESET CLOCK” will be displayed.  
Display of ”RESET CLOCK” will continue until you either  
reset the clock or turn it off.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
5.8 Configuring AutoClean  
Purpose  
The AutoClean feature periodically rinses the cells automatically (see Section 7). If this feature  
was selected during I/O setup (described in 5.3), configure control of AutoClean as described  
below.  
AutoClean configuration consists of specifying the frequency and duration of the rinses.  
ATTENTION  
For AutoClean to operate, the 24-hour clock must be set as described in 5.7.  
The piping and valves required by AutoClean are shown in Section 7.  
There are two modes of operation, timer-initiated and operator-initiated through the front panel  
as described below.  
Timer-initiated operation  
With the 24-hour clock turned ON and the AutoClean feature enabled (and the appropriate piping  
in place), this mode performs unassisted cell cleaning. The AutoClean Time Remaining on-line  
display provides an up-to-date time stamp of the next occurrence of an automatic cell cleaning.  
Once the timer reaches zero, the cell cleaning will occur. Upon completion, the timer will be  
reset and the Time Remaining display will again provide a countdown to the next cleaning. To  
temporarily disable timer-initiated AutoClean, you must either turn off the 24-hour clock (which  
preserves your timer setup), or set the timer to all zeros (see Step 3 in Table 5-8). Unless the  
Time Remaining display is on screen, you will not see the AUTO CLEAN message during a  
scheduled cleaning.  
Operator-initiated operation  
To operate in this mode, you must have the AutoClean Time Remaining display on screen.  
Initiate the AutoClean sequence by pressing the START soft key. Once the START soft key is  
pushed, the time display will go to zero and the message “AUTO CLEAN” will appear on the  
alarm stripe.  
The sequence will stop cleaning if the STOP soft key is pressed or if you leave this display.  
Once stopped, the Analyzer/Controller will wait for the configured resume delay time (to allow  
the cell measurement to stabilize) before coming back on-line. During the resume delay the  
“AUTO CLEAN” message will flash.  
If the PAUSE soft key is used while the AutoClean Time Remaining display is on screen, the  
AutoClean sequence will stop temporarily. If the PAUSE soft key is not pushed again to resume  
the sequence, then the sequence will automatically resume in 20 minutes.  
Once the AutoClean sequence has been completed (including resume delay), the “AUTO  
CLEAN” message disappears and the time remaining clock will be displayed again, showing the  
time remaining until the next scheduled timer-initiated cleaning.  
5-18  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
Hold mode  
During AutoClean operations alarms and outputs are always held. This means that no alarms  
will become active during the automatic operation. If one or more relays are being used for  
alarm annunciation, they will go to the non-alarm state. Analog outputs and discrete control  
relays will be held at their current value or state when the operation starts.  
At the completion of the AutoClean operation the hold will be released and normal operation will  
resume after the configured delay.  
Procedure  
If AutoClean was selected during I/O setup, follow the procedure in Table 5-8 to use the  
displayed screens for configuration when AUTO CLEAN is selected from the ADVANCED  
FEATURES configuration menu.  
Table 5-8 Procedure for Configuring AutoClean  
Step  
1
Screen  
Action  
MENU  
Go to CONFIGURATION | ADVANCED FEATURES |  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
AUTO CLEAN.  
DIAGNOSTICS  
| PREV | NEXT  
CONFIGURATION  
CAT/RETRANSMISSION  
ALARMS  
DISCRETE CONTROL  
ADVANCED FEATURES  
DIAGNOSTICS  
| PREV | NEXT  
ADVANCED FEATURES  
CLOCK SETUP  
COMPUTED VALUE  
AUTO CLEAN  
DIAGNOSTICS  
| PREV | NEXT  
ATTENTION  
24 HOUR CLOCK MUST  
BE SETUP FOR  
2
If you have already set the clock, go to the next screen.  
THIS FEATURE  
If you have not set the clock, press MENU to exit this  
function, then set the clock as described in Sub-Section  
5.7.  
DIAGNOSTICS  
|
| NEXT  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Step  
3
Screen  
Action  
AUTO CLEAN  
Timer setup is used to specify the frequency of the rinses.  
If several rinses are to be performed in a single day:  
TIMER SETUP  
DAYS  
HOURS  
MINUTES  
CURRENT DAY  
3
16  
30  
2
a) specify the number of days as zero  
b) specify the number of hours and minutes in the  
interval between rinses.  
DIAGNOSTICS  
|
| NEXT  
If rinses are to occur at intervals of days:  
a) specify the number of days in the interval  
b) specify the time of day (hours and minutes) the  
rinse should occur.  
In our example, the rinse will occur every third day at  
4:30 PM.  
If you plan to use AutoClean, but initiate rinses manually  
using the real time AutoClean Time Remaining display:  
a) specify the number of days as zero  
b) specify the number of hours and minutes in the  
interval between rinses as zero also.  
IF YOU ENTER ALL ZEROS, NO AUTOMATIC  
RINSES WILL BE PERFORMED.  
Specify where today falls in the AutoClean cycle. For  
example, if the interval for rinses is 3 days and you want  
the first rinse to be tomorrow, specify that today is day 2.  
AUTO CLEAN  
4
Enter the duration of the rinse (in seconds).  
RINSE SETUP  
DURATION  
RESUME DELAY  
Enter the number of seconds the Analyzer/Controller  
should wait after the rinse before resuming normal  
operation.  
180  
60  
DIAGNOSTICS  
| PREV | NEXT  
5-20  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
5.9 Selecting a Computed Value to Be Displayed  
Purpose  
The Analyzer/Controller can calculate and display a computed value. The available values are:  
Cell1  
Cell2  
RATIO: ratio between the values measured by the cells;  
Cell1Cell2  
DIFFERENCE: difference between the values measured by the cells;  
Cell1  
x100  
Cell2  
%PASSAGE: computed percent passage:  
Cell1  
(1−  
)x100  
Cell2  
%REJECTION: computed percent rejection:  
PPMCO2: computed parts per million carbon dioxide (standard range analyzers only)  
Appendix B includes examples of applications using these values.  
ATTENTION  
If a computed value is used, it is significant which cell is considered “Cell 1” and which is  
“Cell 2”. This must be taken into account when wiring the inputs. See 6.3 for details.  
Procedure  
Use the procedure in Table 5-9 to specify a computed value.  
Table 5-9 Procedure for Specifying a Computed Value  
Step  
1
Screen  
Action  
MENU  
Go to CONFIGURATION | ADVANCED FEATURES |  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
COMPUTED VALUE.  
DIAGNOSTICS  
| PREV | NEXT  
CONFIGURATION  
CAT/RETRANSMISSION  
ALARMS  
DISCRETE CONTROL  
ADVANCED FEATURES  
DIAGNOSTICS  
| PREV | NEXT  
ADVANCED FEATURES  
CLOCK SETUP  
COMPUTED VALUE  
AUTO CLEAN  
DIAGNOSTICS  
| PREV | NEXT  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Step  
2
Screen  
Action  
COMPUTED VALUE  
Specify the value to be computed.  
SELECT  
DIFFERENCE  
DIAGNOSTICS  
| NEXT  
|
5-22  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
5.10 Summary of Menu Choices and System Defaults  
Introduction  
Table 5-10 contains the menu choices available (or ranges for numeric values) and system  
defaults for I/O assignments and configuration parameters, as well as for values entered during  
maintenance functions described in Section 10, and the TDS conversion factor entered using the  
calibration menu described in Section 8. Not every choice shown in this table applies to every  
application.  
Table 5-10 Menu Choices and System Defaults  
PARAMETER  
CHOICES or RANGE  
(Default is underlined)  
9782C-S0  
I/O setup: measurement type  
COND/TDS [total dissolved solids]  
RESISTIVITY  
9782C-W0  
COND/TDS  
CONCENTRATION  
9782C-S0  
I/O setup: solution temperature compensation type  
NaCl  
NH3/ACID/ETA  
MORPHOLINE  
NaCl (TDS)  
NH3 (TDS)  
MORPHOLINE(TDS)  
9782C-W0  
NaCl  
HCl  
NaOH  
H2SO4  
NaCl(TDS)  
HCl(TDS)  
NaOH)TDS)  
H2SO4(TDS)  
0.01  
0.1  
I/O setup: cell constant for Cell 1  
1.0  
10.0  
5.0  
25.0  
50.0  
0.01  
0.1  
I/O setup: cell constant for Cell 2  
1.0  
10.0  
5.0  
25.0  
50.0  
N/A  
NO  
YES  
I/O setup: AutoClean enabled  
I/O setup: range status enabled  
NO  
YES  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
PARAMETER  
CHOICES or RANGE  
(Default is underlined)  
NO  
I/O setup: discrete control enabled  
YES  
PFT  
I/O setup: discrete control type for each available relay  
DAT  
O/F [on/off]  
N/A  
ENERGIZE  
I/O setup: relay activation on alarm/control  
DE-ENERGIZE  
I/O setup: display unit of measure (conductivity only)  
standard range models with cell constant 1:  
uS (microSiemens per centimeter)  
uSm (microSiemens per meter)  
standard range models with cell constant of 10,25,or 50:  
uS (microSiemens per centimeter) only  
wide range models: automatially change from uS to mS  
depending on cell constant; no selection available.  
ALM1 [alarm 1] default for Relay 1  
ALM2 [alarm 2] default for Relay 2  
ALM3 [alarm 3] default for Relay 3  
ALM4 [alarm 4] default for Relay 4  
A1/D [alarm 1 or diagnostic]  
A2/D [alarm 2 or diagnostic]  
A3/D [alarm 3 or diagnostic]  
A4/D [alarm 4 or diagnostic]  
A1/2 [alarm 1 or 2]  
I/O setup: relay alarm assignments  
A3/4 [alarm 3 or 4]  
12/D [alarm 1 or 2 or diagnostic]  
AA/D [any alarm or diagnostic]  
234D [alarm 2 or 3 or 4 or diagnostic]  
RNGD [range switch or diagnostic]  
AA [any alarm]  
RNGS [range switch]  
C1 [Cell 1 value]  
I/O setup: analog output basis [unless DE is enabled]  
C2 [Cell 2 value]  
TMP1 [temperature value from Cell 1]  
TMP2 [temperature value from Cell 2]  
Cell1  
Cell2  
RATO [computed ratio:  
]
Cell1  
Cell2  
(1−  
)x100  
%REJ [computed percent rejection:  
]
Cell1  
x100  
Cell2  
%PSG [computed percent passage:  
]
Cell1Cell2  
DIFF [computed difference:  
]
CO2 [computed parts per million carbon dioxide -  
standard range analyzers only]  
LINFIXEDRNG [linear fixed range] always selected for  
computed value  
configuration: retransmission output scaling  
LINAUTORNG [linear auto range]  
2DECADELOG  
BI-LINEAR  
5-24  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
I/O Setup and System Configuration  
PARAMETER  
CHOICES or RANGE  
(Default is underlined)  
range depends on measurement type and cell constant;  
default is value of lower range limit  
configuration: retransmission 0%  
configuration: retransmission 100%  
range depends on measurement type and cell constant;  
default is value of upper range limit  
configuration: CAT control setpoint value  
configuration: alarm setpoint value  
configuration: configuration: On/Off control cycle timer  
setpoint value  
configuration: PFT control setpoint value  
configuration: DAT control setpoint value  
range depends on measurement type and cell constant;  
default is value of upper range limit  
configuration: CAT control proportional band limit  
configuration: alarm deadband  
configuration: On/Off control cycle timer deadband  
configuration: PFT control proportional band limit  
configuration: DAT control proportional band limit  
0 to 100 seconds, default = 0  
configuration: alarm delay  
C1 [Cell 1 value]  
configuration: process variable or computed value to  
C2 [Cell 2 value]  
be used as basis for discrete control output  
TMP1 [temperature value from Cell 1]  
TMP2 [temperature value from Cell 2]  
T1/2 [temperature 1 and/or 2 - only available for  
configuration: process variable or computed value  
alarmed  
alarming]  
Cell1  
Cell2  
RATO [computed ratio:  
]
Cell1  
Cell2  
(1−  
)x100  
%REJ [computed percent rejection:  
]
Cell1  
x100  
Cell2  
%PSG [computed percent passage:  
]
Cell1Cell2  
DIFF [computed difference:  
]
CO2 [computed parts per million carbon dioxide -  
9782C-S0 only]  
HIGH  
LOW  
configuration: alarm action  
0 to 1000 seconds, default = 60  
configuration: On/Off control cycle timer period  
configuration: On/Off control cycle timer “on” time  
configuration: PFT control maximum pulse frequency  
configuration: DAT control cycle period  
0 to 100% at 1% resolution, default = 100  
1 to 200 pulses per minute, default = 60  
0 to 200 seconds, default = 60  
OFF  
configuration: 24-hour clock set mode  
ON W/ RESET  
ON W/O RESET  
0 to 24, default = 1  
0 to 59, default = 0  
0 to 28, default = 0  
0 to 24, default = 1  
0 to 59, default = 0  
configuration: 24-hour clock set hour  
configuration: 24-hour clock set minute  
configuration: AutoClean timer set day  
configuration: AutoClean timer set hour  
configuration: AutoClean timer set minutes  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
5-25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
PARAMETER  
CHOICES or RANGE  
(Default is underlined)  
0 to 28, default = 0  
configuration: AutoClean current day  
1 to 1999 seconds, default = 60  
configuration: AutoClean rinse duration  
configuration: AutoClean resume delay  
NONE  
configuration: computed value selection  
Cell1  
Cell2  
RATIO [computed ratio:  
]
Cell1Cell2  
DIFFERENCE [computed difference:  
]
Cell1  
Cell2  
x100  
%PASSAGE [computed percent passage:  
%REJECTION [computed percent rejection  
]
Cell1  
(1−  
)x100  
Cell2  
:
]
PPMCO2 [computed parts per million carbon dioxide -  
9782C-S0 only]  
LOW  
HIGH  
maintenance: display resolution  
maintenance: security enabled  
maintenance: password  
YES  
NO  
0000 to 9999; default = 0000  
12  
14  
16  
18  
maintenance: lead resistance compensation wire gauge  
(AWG)  
wide range models only  
1 to 1500 feet default = 1  
maintenance: lead resistance compensation length  
wide range models only  
standard range models with cell constant of 1, 10, 25, or  
50:  
calibration: TDS conversion factor  
0.01 to 1.999; default = 0.5 ppm  
standard range models with cell constant of 0.01 or 0.1:  
10 to 1999; default = 500 ppb  
wide range models for all cell constants:  
0.01 to 1.999; default = 0.5 ppm  
5-26  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Input and Output Wiring  
6. Input and Output Wiring  
6.1 Overview  
Introduction  
This section contains instructions for installing input and output wiring for the  
Analyzer/Controller.  
We recommend that you wait to install input and output wiring until after I/O setup (see  
Section 5). During I/O setup the software will determine for you which relay to use for each  
feature.  
ATTENTION  
If you plan to use a computed value such as the difference between the cell readings, it matters  
which cell is “Cell 1” and which is “Cell 2”. See 6.4.3 for details.  
ATTENTION  
For successful measurement in pure water applications where plastic piping is used, you may  
have to provide an earth ground for the cell. If the cell constant is 0.01, 0.1, or 1.0, run a wire  
from the screw terminal (on the Analyzer/Controller ) to which the black wire in the cell lead is  
attached to the lower card retainer bracket screw. (See Note 4 in Figure 6-1.)  
Do not ground cells having cell constants of 10, 25, or 50.  
What’s in this section?  
The topics in this section are listed below.  
Topic  
See Page  
6-1  
6.1 Overview  
6.2 General Wiring Practices for I/O Wiring  
6.3 Wiring Analog Inputs and Outputs  
6.4 Wiring Relays  
6-2  
6-4  
6-7  
WARNING  
While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not open  
the case while the unit is powered. Disconnect power before installing I/O wiring. More than one  
switch may be required to disconnect power.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
6.2 General Wiring Practices for I/O Wiring  
WARNING  
Wiring should be performed by qualified personnel only.  
Safety precaution  
WARNING  
A disconnect switch must be installed to break all current carrying conductors. Turn off power  
before working on conductors. Failure to observe this precaution may result in serious personal  
injury.  
WARNING  
An external disconnect switch is required for any hazardous voltage connections to the relay  
outputs.  
Avoid damage to components  
ATTENTION  
This equipment contains devices that can be damaged by electrostatic discharge (ESD).  
The damage incurred may not cause the device to fail completely, but may cause early failure.  
Therefore, it is imperative that assemblies containing static sensitive devices be carried in  
conductive plastic bags. When adjusting or performing any work on such assemblies,  
grounded work stations and wrist straps must be used. If soldering irons are used, they must  
also be grounded.  
A grounded work station is any conductive or metallic surface connected to an earth ground,  
such as a water pipe, with a 1/2 to 1 megohm resistor in series with the ground connection. The  
purpose of the resistor is to current limit an electrostatic discharge and to prevent any shock  
hazard to the operator. The steps indicated above must be followed to prevent damage and/or  
degradation, which may be induced by ESD, to static sensitive devices.  
Immunity Compliance  
In applications where either the power, input or output wiring are subject to electromagnetic  
disturbances, shielding techniques will be required. Grounded metal conduit with conductive  
conduit fittings is recommended.  
Conform to code  
Instrument wiring should conform to national and local electrical codes.  
6-2  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Input and Output Wiring  
Recommended maximum wire size  
Table 6-1 Recommended Maximum Wire Size  
Gage Number  
mm2  
Description  
power and relays  
14  
see Parts List  
18  
2.081  
0.823  
0.823  
inputs*  
isolated outputs  
* Automatic compensation for input lead resistance is available. See Section 10 for details.  
Cables for locations without interference  
For steel (metal) conduit or open wiring away from any sources of interference, use  
recommended Honeywell cables listed in the parts list in Section 11.  
Shielded wiring for locations with interference  
In applications where plastic conduit or open wire trays are used, shielded 6-conductor 22 gage  
(0.326 mm2) or heavier signal input wiring is required.  
Avoiding interference  
Instrument wiring is considered Level 1, per section 6.3 of IEEE STD. 518 for plant facilities  
layout and instrumentation application. Level 1 wiring must not be run close to higher level  
signals such as power lines or drive signals for phase fired SCR systems, etc. Unprotected input  
wiring in high electrical noise environments is subject to electromagnetic, electrostatic, and radio  
frequency interference pickup of sufficient magnitude to overload input filters. The best  
instrument performance is obtained by keeping the interfering signals out of the instruments  
altogether by using proper wiring practices.  
References  
Refer to the following when wiring the unit.  
IEEE STD. 518, Guide for the Installation of Electrical Equipment to Minimize Electrical  
Noise Inputs to Controllers from External Sources.  
Appropriate wiring diagram supplied with cells.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
6.3 Wiring Analog Inputs and Outputs  
Introduction  
Every 9782 Analyzer/Controller requires input signals from the cells.  
In addition, some 9782 models provide from one to three analog output signals (current and/or  
voltage) that can be used to retransmit process variables, or to provide Current Adjusting Type  
control (see 1.5.3).  
Wiring these analog inputs and outputs is described here. Wiring relay outputs is described in  
6.4.  
Cell identification significant to computed value  
The 9782 can compute a value (choices listed below), then display the value, and, if desired, use  
it as the basis for control or alarming. (Sample applications are shown in Appendix B.) Each of  
these computed values is calculated based on the process variable value measured by both cells.  
Therefore, it matters which cell is considered “Cell 1” and which is “Cell 2”. The  
Analyzer/Controller determines the cell’s identity based upon which terminals receive the cell’s  
input (see Figure 6-1). If you have set up the 9782’s software to use a computed value, be  
sure to wire the cell leads correctly.  
The available computed values are:  
Cell1  
Cell2  
RATIO: ratio between the values measured by the cells;  
Cell1Cell2  
DIFFERENCE: difference between the values measured by the cells;  
Cell1  
x100  
Cell2  
%PASSAGE: computed percent passage:  
Cell1  
(1−  
)x100  
Cell2  
%REJECTION: computed percent rejection:  
PPMCO2: computed parts per million carbon dioxide (9782C-S0 only)  
Procedure  
Follow the procedure in Table 6-2 to install analog I/O wiring.  
WARNING  
While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not open  
the case while the unit is powered. Disconnect power before installing I/O wiring. More than one  
switch may be required to disconnect power.  
6-4  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Input and Output Wiring  
Table 6-2 Procedure for Installing Analog I/O Wiring  
Step  
1
Action  
Go to I/O SETUP to view the displays showing analog input, relay, and analog output use.  
Note the assignments shown. You must wire the unit to match these assignments in order  
for the unit to work as expected. (See Section 5.)  
2
3
Turn off the power to the Analyzer/Controller. More than one switch may be required  
to remove power.  
With the power off, open the case:  
Grasp the bottom center portion of the front bezel and pull it downward and toward you  
slightly to disengage the bottom of the bezel from the edge of the case.  
Lift the bezel gently to disengage it from the top edge of the case.  
Swing the bezel to the left. (The bezel and display assembly is mounted on pivot arms.)  
The safety cover with its warning label will be visible.  
4
5
With the power off remove the safety cover by removing the single screw holding it in  
place.  
Connect the inputs from the cells to the terminals in accordance with the system setup  
assignments. Refer to the wiring diagram provided with the cells and to Figure 6-1*.  
* There are 2 resistors, one for cell 2 input and one for cell 2 thermistor input, which are installed in every unit at  
the factory. These resistors are required if only cell 1 input is being utilized. If cell 2 input is being utilized, remove  
these resistors from the unit. These resistors can be saved and used as Precision Check Resistors. See  
Appendix D for details.  
If analog outputs are to be used, connect the outputs from the Analyzer/Controller terminals  
in accordance with the system setup assignments. Refer to the wiring diagrams provided  
with the field devices receiving the signals and to Figures 6-1.  
6
If the relay outputs are to be used, leave the unit open and powered down. See 6.4.  
If the relays will not be used, replace the safety cover and secure it with the screw.  
Close the case and power up the unit. Do not apply power with the case open.  
7
8
9
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Relay Terminals  
(optional)  
Case (as viewed from front)  
NO  
Note 3  
Relay 3  
Note 6  
COM  
NC  
NO  
SH  
G
COM  
NC  
NO  
COM  
NC  
Relay 4  
Note 6  
No. 2 Temp.  
Compensator  
NO  
R
G
COM  
NC  
I(+)  
I(-)  
+10V 2000 min  
VCom  
4-20 mA 600 min  
No. 1 Temp.  
Compensator  
R
W
+1V 200 min  
Cell 2  
Electrodes  
NC  
1
2
K
CO M  
Relay 2  
Note 1  
W
3
4
5
6
7
8
NO  
NC  
+
Cell 1  
Electrodes  
mA  
-
Relay 1  
Note 1  
K
CO M  
NO  
+10  
VC  
+1  
L1 (K)  
L2 /N  
120V/240V  
15A  
50-60 Hz  
Note 5  
NOTES:  
Note 4  
1. Relay 1 & 2 contacts max. ratings  
(Resistive Load)  
9
Current: 2A AC or DC  
Voltage: 250 VAC, 220 VDC  
Switching Power: 125VA, AC: 60 W, DC.  
2. See Cell Connection Diagram or directions  
for wiring restrictions.  
Power &  
Voltage & Current  
Output Terminals  
Relay Terminals  
3. Coaxial cable used with greater than 20 ft.  
cell/Analyzer seperation with Standard  
Range instruments only.  
Two-Cell Cat. 9782 with Output/relays  
4. For pure water samples in plastic piping,  
ground the black cell electrode lead near  
the cell. Alternately, connect to the 9782  
ground screw as shown dotted. DO NOT  
ground 10, 25 or 50 constant cells.  
5. Analyzer is factory set for 120V operation.  
For 240V operation, see "Power  
Connections" in this manual.  
Case (as viewed from front)  
Output Board  
Connections  
(4-20 mA)  
Note 3  
O UT 1 -  
1
SH  
O UT 1 +  
2
No. 2 Temp.  
Compensator  
6. Relay 3 & 4 contact ratings 3A @ 120V AC/  
28 VDC resistive.  
G
O UT 2 -  
3
4 O UT 2 +  
R
O UT 3 -  
5
1
NC  
No. 1 Temp.  
Compensator  
G
6 O UT 3 +  
Relay 2  
Com  
2
R
Note 1  
NO  
3
Cell 2  
W
Electrodes  
4
NC  
K
Color Code  
5
Relay 1  
Com  
Cell 1  
Electrodes  
W
Note 1  
NO  
6
K- black  
Y- yellow  
K
7
8
9
L1(K)  
L2/N  
120V/240V 15A  
50-60 Hz  
Note 5  
W- white A- gray  
Note 4  
G- green  
R- red  
V- violet  
P- pink  
S- slate  
B- blue  
Power &  
Relay Terminals  
N- brown T- tan  
O- orange  
Two-Cell Cat. 9782 - MultiOptions  
a/n 23177  
Figure 6-1 Terminal Connections  
6-6  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Input and Output Wiring  
6.4 Wiring Relays  
6.4.1 Installation  
Introduction  
Every 9782 model includes two relay outputs that can be used with special features such as  
AutoClean and output range status indication, or to perform On/Off, Pulse Frequency Type or  
Duration Adjusting Type control, or for alarm annunication. In additon to these two standard  
relays, two more are available as an option.  
The Analyzer/Controller can be programmed for these relays to de-energize on alarm, or to  
energize on alarm. Use the maintenance menu to specify relay action. (see 10.11.)  
If set to de-energize on alarm, this means that when an alarm occurs (or the discrete control point  
becomes active), the relay coil will be de-energized. The NC contacts will then be closed and the  
NO contacts will be open. Conversely, during normal non-alarm operation (or when the control  
point is not active) the NC contacts will be open, and the NO contacts will be closed.  
If de-energize on alarm is selected, a power loss will force all relays to the same position as an  
alarm condition.  
The relay contact maximum ratings are in Table 6-3.  
Table 6-3 Relay Contact Maximum Ratings  
0.6 amps at 120/240 Vdc, 0.6 amps at 110 Vdc,  
2 amps at 30 Vdc  
Current, Standard:  
3A, 120 Vac or 28 Vdc  
120/240 Vac  
Current, Optional:  
Voltage, Standard:  
ac - 125 VA resistive  
dc - 60 watts resistive  
Switching Power, Standard:  
CAUTION  
Alarm circuits are not internally fused in the Analyzer/Controller. Provision for fuses in external  
circuits is recommended.  
Procedure  
Follow the procedure in Table 6-3 to install relay output wiring.  
If the RC arc suppression circuits in the Analyzer/Controller interfere with other devices at your  
site, refer to 6.4.2 for instructions for disconnecting the suppression circuits.  
If a relay is being used to indicate automatic output range change, see 6.4.3 for guidance in  
wiring a device to the relay.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
WARNING  
While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not  
open the case while the unit is powered. More than one switch may be required to  
disconnect power.  
Table 6-4 Procedure for Wiring Relays  
Step  
1
Action  
If you have not already noted the relay assignments made during I/IO setup, go to I/O  
SETUP on the main menu to view the display showing relay output use. Note the  
assignments shown. You must wire the unit to match these assignments in order for the  
unit to work as expected. (See Section 5.)  
2
3
Turn off the power to the Analyzer/Controller. More than one switch may be required  
to remove power.  
With the power off, open the case:  
Grasp the bottom center portion of the front bezel and pull it downward and toward you  
slightly to disengage the bottom of the bezel from the edge of the case.  
Lift the bezel gently to disengage it from the top edge of the case.  
Swing the bezel to the left. (The bezel and display assembly is mounted on pivot arms.)  
The safety cover with its warning label will be visible.  
4
5
With the power off remove the safety cover by removing the single screw holding it in  
place.  
Wire the relays in accordance with the system setup assignments. Refer to the wiring  
diagram provided with the external device receiving the output and to Figure 6-1.  
WARNING  
An external disconnect switch is required for any hazardous voltage connections to the relay outputs.  
Replace the safety cover and secure it with the screw.  
6
7
Close the case and power up the unit. Do not apply power until the case is closed.  
6-8  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Input and Output Wiring  
6.4.2 Disconnecting RC Arc Suppression Circuits  
Introduction  
Relay contacts are protected by RC arc suppression circuits. Alternating current leakage through  
these circuits may be sufficient to trigger very light ac loads such as neon bulbs, PLC logic  
inputs, continuity testers, etc. If this leakage cannot be tolerated, the suppression circuits may be  
disconnected as described below.  
Procedure  
Follow the procedure in Table 6-5 to disconnect the RC arc suppression circuits.  
Table 6-5 Procedure for Disconnecting RC Arc Suppression Circuits  
Step  
Action  
1
Turn off the power to the Analyzer/Controller. More than one switch may be required  
to remove power.  
2
With the power off, open the case:  
Grasp the bottom center portion of the front bezel and pull it downward and toward you  
slightly to disengage the bottom of the bezel from the edge of the case.  
Lift the bezel gently to disengage it from the top edge of the case.  
Swing the bezel to the left. (The bezel and display assembly is mounted on pivot arms.)  
The safety cover with its warning label will be visible.  
3
With the power off remove the safety cover by removing the single screw holding it in  
place.  
Disconnect and tag all wiring from the terminal boards.  
4
5
Remove the circuit card retainer clamp at the top of the case. Then slide all the circuit cards  
half way out of the case.  
On the right-most vertical printed circuit card (power supply) remove 330 ohm resistors R29  
and R30 for Relay 1, and R31 and R32 for Relay 2 from the upper corner of the card.  
6
7
If the auxiliary relays are installed, also remove 330 ohm resistors R39, R40, R44 and R45  
from those optional assemblies (see Figure 6-1).  
Reconnect all wiring removed in Step 4.  
8
9
Replace the safety cover and secure it with the screw.  
10  
Close the case and power up the unit. Do not apply power until the case is closed.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
6.4.3 Wiring a Device to the Relay Indicating Output Range Status  
Introduction  
If the automatic output range change feature is selected during I/O setup, one relay is dedicated  
to this function. The relay can be used to control an indicator, or to switch the output to a second  
channel of a multipoint recorder when the range change occurs (see Figure 6-2). Check the  
current relay assignments as described in Section 5 to determine whether a relay was assigned to  
this feature.  
This feature closes contacts NC-COM when the output is scaled to the low range (Range 1 in  
Figure 6-2) and closes contacts NO-COM when scaled to the high range (Range 2 in Figure  
6-2).  
Interconnections  
The range change capability illustrated in Figure 6-2 is accomplished with interconnections  
shown in Figure 6-3. Similar range identification could be obtained using a 2-pen or event-pen  
recorder.  
Point 1/Terminals 1 - low range (0-1µS/cm)  
Point 2/Terminals 2 - high range (0-10 µS/cm)  
Resistors “R” across the recorder inputs are needed to ensure that the unused point will remain at  
the low end of scale. The resistance value can be anything from 500 ohms to 1 Mohm.  
(simulated  
condenser  
leak)  
Range 1  
Range 2  
0
0
.2  
2
.4  
4
.6  
6
.8  
8
1.0  
10  
a/n 23178  
Figure 6-2 Sample Multipoint Recorder Chart Illustrating Auto Output Range  
Change for Steam Condensate Monitoring  
6-10  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Input and Output Wiring  
Multi-point Recorder  
Input Terminal Board  
6
4
2
6
5
3
1
5
4
9782 Conductivity Analyzer  
3
2
R
R
1
NO  
NC  
1
2
3
4
+
mA  
-
5
6
+10  
VC  
7
+1  
8
9
where R = 250 ohms  
a/n 23179  
Figure 6-3 Interconnections between 9782 and Recorder with Range Identification  
by Point Change  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
6-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
6-12  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
AutoClean Theory and Piping  
7. AutoClean Theory and Piping  
7.1 Overview  
Introduction  
The AutoClean feature periodically rinses the cells automatically as described in this section. To  
take advantage of this feature you must:  
Select it during I/O setup (see Section 5).  
Set the clock, and specify the frequency and duration of the rinse (see Section 5).  
Wire the relay assigned to this operation during I/O setup to operate the necessary valves.  
(See Section 6 for relay wiring.)  
Install piping and valves as diagrammed in this section.  
What’s in this section?  
The topics in this section are listed below.  
Topic  
See Page  
7-1  
7.1 Overview  
7.2 AutoClean Sequence and Piping  
7-2  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
7.2 AutoClean Sequence and Piping  
Rinse sequence  
The AutoClean operation occurs at the configured intervals. (It can also be initiated by the  
operator; see Sub-Section 5.8 for details.) The sequence is described below.  
1
All alarm action is held at existing levels. The outputs are also held throughout the  
AutoClean operation.  
Pressing the DISPLAY key will call up a special display that shows how much time  
remains until the start of the next scheduled AutoClean sequence. While the cleaning  
operation is taking place, “AUTO CLEAN” is displayed on the alarm stripe on the Time  
Remaining display. The clock will show 00:00:00.  
2
3
4
Relay 1 activates 3-way solenoid valve S1 (see Figure 7-1) to direct rinsing fluid to the  
cells for the configured rinse duration (1 to 1999 seconds). If the measured sample is  
normally returned to the process but quantities of rinsing fluid cannot be tolerated there,  
use an additional 3-way solenoid valve S4. It is activated simultaneously with S1 to divert  
the discharge to drain.  
At the end of the configured rinse time Relay 1 de-activates solenoid valve S1 (and S4, if  
used). After the configured delay period (1 to 1999 seconds) the Analyzer/Controller  
resumes sampling the process. (During the resume delay the “AUTO CLEAN” message  
flashes.) Note that even with S4 for diversion, one system volume of washing fluid will  
pass to the process at this point.  
The “AUTO CLEAN” message is cleared.  
Note that the operator can make the operation pause using the “soft key” on the special  
AutoClean Time Remaining display. If the operator does not remove the pause by pressing the  
PAUSE soft key again, the Analyzer/Controller will resume normal operation after 20 minutes.  
7-2  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
AutoClean Theory and Piping  
rinse/cleaning  
solution  
9782  
to process  
cell  
S4  
to drain  
S1  
process  
sample  
Items outside this area  
provided byuser  
Automatic Wash Setup  
Outputs & Alarms Held  
Rinse Duration (seconds)  
Relay 1 Activated  
Resume DelayTime (seconds)  
NormalOperation  
0.222"  
1.709"  
0.588"  
TIME  
Interval Between Rinses  
(daysor hours)  
AutoClean Operation  
a/n 23185  
Figure 7-1 Automatic Cell Wash Setup  
Select valves and fittings with appropriate pressure ratings  
Make the process connections as shown in Figure 7-1. Be sure that valves and fittings (S1) have  
sufficient pressure ratings to withstand pressure peaks which will occur when process flow is  
blocked.  
Minimize liquid volume in system  
Keep pipe sizes small and couplings close to minimize the liquid volume in the system. Smaller  
volumes require less time to rinse.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
7-4  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
8. Calibration  
8.1 Overview  
Introduction  
Each type of cell has an associated cell constant entered during I/O setup (see Section 5). This  
number is part of the cell model number. However, for greater precision, every Honeywell cell is  
individually tested at the factory, and a calibration factor unique to that cell is determined. The  
cal factor for a cell can be found on the plastic tag hanging from the cell lead wires. Instructions  
for entering this cell cal factor are in 8.3.  
For some conductivity applications even greater accuracy is required. For those applications it is  
possible to perform a calibration trim procedure. The Analyzer/Controller’s reading can be  
adjusted while the associated cell is measuring a reference solution of known conductivity, as  
described in 8.4. The same procedure can be used to adjust the Analyzer/Controller’s reading  
while the cell is in the process, if a reference instrument is used to determine the conductivity of  
the process. In this case the process fluid becomes the “reference solution”.  
Calibration trim can be removed as described in 8.5  
For accurate measurement of total dissolved solids (TDS) a conversion factor is entered for each  
cell as described in 8.6.  
ATTENTION  
Any time a cold reset is performed, the calibration trim and TDS values will be removed.  
The cal factor will be reset to match the cell constant.  
What’s in this section?  
The topics in this section are listed below.  
Topic  
See Page  
8-1  
8.1 Overview  
8.2 Recommendations for Successful Measurement and Calibration  
8.3 Entering the Cal Factor for Each Cell  
8.4 Performing Calibration Trim  
8-3  
8-4  
8-5  
8.5 Removing Calibration Trim Values  
8.6 Determining and Entering the TDS Conversion Factor  
8-7  
8-8  
Calibration menu customized for different configurations  
The choices available on the menu displayed when CALIBRATION is selected from the main  
menu depend on the choices made for solution temperation compensation during I/O setup (see  
Step 13 in Table 5-1).  
If neither cell’s solution temperature setup uses total dissolved solids (TDS), then the calibration  
menu contains the following choices:  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
8-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
CELL CAL FACTOR  
CAL TRIM C1  
CAL TIRM C2  
REMOVE CAL TRIM C1  
REMOVE CAL TRIM C2  
If the Cell 1 solution temperature compensation is set up without TDS, and Cell 2 with TDS,  
then the calibration menu will be:  
CELL CAL FACTOR  
CAL TRIM C1  
TDS CVSN FACTOR C2  
REMOVE CAL TRIM C1  
REMOVE CAL TRIM C2  
If both cells’ solution temperature compensation is set up with TDS, then the calibration menu  
will be:  
CELL CAL FACTOR  
TDS CVSN FACTOR C1  
TDS CVSN FACTOR C2  
REMOVE CAL TRIM C1  
REMOVE CAL TRIM C2  
Therefore, the screens on your unit may not match exactly the calibration menus used in the  
procedure tables in this section.  
8-2  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
8.2 Recommendations for Successful Measurement and Calibration  
Selection and care of cell essential  
Successful measurements and calibration depend upon selection and care of the cells. Always  
prepare cells and their mountings in accordance with the instructions supplied with them,  
observing temperature, pressure and flow limitations.  
Recommendations  
Note the following recommendations:  
For most accurate temperature measurement and compensation, insulate the outer body of the  
cell to minimize the effect of ambient conditions on process temperature measurement.  
Rinse the cells thoroughly with de-ionized water before immersing in a reference solution.  
Always deactivate control or alarm circuits before removing cells from the process.  
Do calibration trim with a reference solution which is at about the same temperature and  
conductivity as the process solution.  
Inspect and, if necessary, clean the cells periodically according to experience and conditions.  
ATTENTION  
For successful measurement in pure water applications where plastic piping is used, you may  
have to provide an earth ground for the cell. If the cell constant is 0.01, 0.1, or 1.0, run a wire  
from the black electrode terminal of the cell to the lower card retainer bracket screw. Do not  
ground cells having cell constants of 10, 25, or 50.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
8-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
8.3 Entering the Cal Factor for Each Cell  
Introduction  
Each type of cell has an associated cell constant; this number is part of the cell model number.  
The constant for each cell is entered during I/O setup. However, for greater precision, every  
Honeywell cell is individually tested at the factory, and a calibration factor unique to that cell is  
determined. The cal factor for a cell can be found on the plastic tag hanging from the cell lead  
wires.  
ATTENTION  
Any time you replace a cell, even with another cell of the same type, enter the cal factor for the  
replacement cell.  
Procedure  
To enter the cal factor for each cell, follow the procedure in Table 8-1.  
Table 8-1 Procedure for Entering Cal Factor  
Step  
1
Screen  
Action  
MENU  
CONFIGURATION  
CALIBRATION  
Go to CALIBRATION | CELL CAL FACTOR.  
MAINTENANCE  
I/O SETUP  
The cell constant entered for each cell during I/O setup will  
be displayed.  
DIAGNOSTICS  
|
|
To enter the unique cal factor for each cell, use the  
indicated function key to go into edit mode.  
CALIBRATION  
CELL CAL FACTOR  
CAL TRIM 1  
CAL TRIM 2  
REMOVE CAL TRIM C1  
REMOVE CAL TRIM C2  
DIAGNOSTICS  
|
|
CELL CONSTANTS  
2
3
The current cell constant for each cell will be displayed.  
This information is read-only.  
C1  
C2  
10.0  
10.0  
To change the values, use the function key labeled  
“NEXT”. To exit, use the MENU key.  
DIAGNOSTICS  
|
| NEXT  
EDIT CELL CAL FACTOR  
C1 10.101  
C2 09.955  
For each cell, enter the cal factor found on the plastic tag  
hanging from the cell leads.  
DIAGNOSTICS  
DPP | <-- | -->  
8-4  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
8.4 Performing Calibration Trim  
Introduction  
For most applications entering the cal factor for each cell will achieve satisfactory system  
performance. However, it is possible to perform a calibration trim procedure in which the  
Analyzer/Controller and cell combination are used to measure a reference solution of known  
conductivity; the reading of the Analyzer/Controller is adjusted to match.  
The same procedure can be used to adjust the Analyzer/Controller’s reading while the cell is in  
the process, if a reference instrument is used to determine the process conductivity. In this case,  
the process fluid becomes the “reference solution”.  
Materials and procedure  
To perform calibration trim using a standard reference solution, follow the instructions in Table  
8-3.  
Materials required are:  
A reference solution of known conductivity near the point of interest, with the temperature  
controlled (or measured and compensated) to within ± 1 °C. Conductivities of potassium  
chloride solutions are provided in Table 8-2. Solutions must be prepared with high-purity  
de-ionized, CO2-free water, and dried potassium chloride.  
A container for the reference solution, large enough to immerse the cell to measuring depth.  
De-ionized water to rinse the cell.  
Table 8-2 Conductivity of Potassium Chloride Solutions at 25 °C  
Concentration M*  
Conductivity (microsiemens  
per cm)  
0.001  
0.005  
0.01  
147.0  
717.8  
1,413  
2,767  
6,668  
0.02  
0.05  
* M = Molarity; 1M = 74.555g potassium chloride per liter of solution  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
8-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Table 8-3 Procedure for Performing Calibration Trim Using a Reference Solution  
Step  
1
Screen  
Action  
MENU  
CONFIGURATION  
CALIBRATION  
Prepare container of the reference solution.  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
|
|
CALIBRATION  
CELL CAL FACTOR  
CAL TRIM 1  
2
3
Go to CALIBRATION | CAL TRIM C 1.  
CAL TRIM 2  
REMOVE CAL TRIM C1  
REMOVE CAL TRIM C2  
DIAGNOSTICS  
|
|
CAL TRIM  
PUT C1 IN  
REFERENCE SOLUTION  
ATTENTION  
WAIT FOR STABLE  
Remove the cell from the process.  
Rinse the cell thoroughly with de-ionized water.  
Immerse it in the reference solution.  
Go to the next screen.  
READING!  
DIAGNOSTICS  
| PREV | NEXT  
CAL TRIM C1  
4
The display will show the conductivity of the reference  
solution as measured by the cell and Analyzer/Controller  
system.  
147.00 µS  
DIAGNOSTICS  
|
Once the reading is stable, change the value to match the  
actual conductivity of the reference solution at its current  
temperature.  
|
Press ENTER to save the calibration trim value.  
If the calibration trim adjustment is successful, the  
calibration menu will again be displayed.  
Return the cell to the process and go to Step 6.  
CAL TRIM C1  
147.00 µS  
5
6
If the attempted calibration trim is greater than the  
allowable adjustment limit, the operation will not be  
successful. The CAL TRIM display will remain on the  
screen and an error message will be displayed on the  
alarm stripe.  
CAL ADJ FAILED  
|
|
Check the cell and replace if necessary.  
After successful calibration, repeat the operation for the  
other cell. When calibration trim has been completed for  
both cells, return to the main menu.  
8-6  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
8.5 Removing Calibration Trim Values  
Introduction  
If the calibration is suspect, or if the system operation is being checked using a reference resistor  
as described in Appendix D, remove the calibration trim. The Analyzer/Controller will go back  
to using the cal factors entered as described in 8.2.  
Procedure  
To remove the calibration trim value for each cell, follow the procedure in Table 8-5.  
Table 8-4 Procedure for Removing Calibration Trim Values  
Step  
1
Screen  
Action  
MENU  
CONFIGURATION  
Go to CALIBRATION | REMOVE CAL TRIM C1.  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
|
|
CALIBRATION  
CELL CAL FACTOR  
CAL TRIM 1  
CAL TRIM 2  
REMOVE CAL TRIM C1  
REMOVE CAL TRIM C2  
DIAGNOSTICS  
|
|
REMOVE CAL TRIM  
ATTENTION!  
THIS FUNCTION WILL  
CANCEL ANY PREVIOUS  
CALIBRATIONS  
2
3
A calibration trim message will be displayed.  
Go to the next screen.  
DIAGNOSTICS  
|
|
REMOVE CAL TRIM  
C1 NO  
To specify that you want to remove the trim for the cell,  
change the NO to YES.  
To execute the removal, press ENTER.  
DIAGNOSTICS  
|
|
To exit the screen without removing the calibration trim  
values, press the MENU key.  
Select “REMOVE CAL TRIM C2” from the Calibration  
Menu and repeat the procedure to remove the calibration  
trim for Cell 2.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
8-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
8.6 Determining and Entering the TDS Conversion Factor  
Introduction  
The 9782 always measures conductivity. However, the process value can be displayed in terms  
of total dissolved solids (TDS). If a TDS solution temperature compensation type was specified  
during I/O setup, then the calibration menu will contain items used to enter the TDS conversion  
factor for each cell.  
Determining TDS conversion factor  
To determine the TDS conversion factor, it is first necessary to establish the total dissolved  
solids in a representative sample of the process. The formal determination of TDS is a laboratory  
standard method performed on a weighed grab sample of the process fluid. To summarize how a  
obtain a TDS value:  
Suspended solids, if present, are filtered out.  
All water is evaporated.  
The residue is dried and weighed.  
The result is divided by the original sample weight to obtain ppm TDS.  
For detailed guidance in determining the official TDS, see “Standard Methods for the  
Examination of Water and Wastewater,” jointly published by the American Public Health  
Association, American Water Works Association and Water Pollution Control Federation,  
Washington, DC.  
To determine the conversion factor needed by the 9782 Analyzer/Controller, first use the  
laboratory procedure summarized above to give an official TDS value. Next divide the TDS  
value by the conductivity of the sample to yield the conversion factor for that particular process  
fluid. The conversion factor is then entered into the 9782 to normalize the TDS readout.  
With power plant cation conductivity measurements, ion chromatography results may be used to  
establish the conversion factor for readout in ppb chloride or sulfate ion. Nominal values are 83  
ppb per µS/cm for chloride ion and 111 ppb per µS/cm for sulfate ion. The 9782 does not  
provide temperature compensation in TDS for chloride or sulfate ions.  
Out-of range-values forced to closest limit  
As long as the entered TDS value is within the acceptable limits for a given cell constant, the  
value is accepted by the Analyzer/Controller. If a value is outside the accepted range, the unit  
will not display an error message; instead it will force the value to either the high or low limit of  
the range of the cell constant. See Table 5-10 for TDS conversion factor defaults.  
Calibrate the Analyzer/Controller before entering TDS conversion factor  
If you intend to enter a cal factor or use calibration trim, do so before entering the TDS  
conversion factor as described here.  
If you use calibration trim, first set the solution temperature compensation in I/O setup to the  
non-TDS choice for your process. For example, if you plan to use “NaCl (TDS)” set the solution  
temperature compensation type to “NaCl” temporarily for calibration purposes. (Solution  
temperature compensation type is the one I/O setup parameter that can be changed without  
triggering a cold reset.)  
8-8  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration  
Next, perform the calibration. Once calibration has been completed, go back to I/O setup and set  
the solution temperature compensation type to the TDS choice, for example “NaCl (TDS)”.  
At this point you are ready to enter the TDS conversion factor as described in Table 8-5.  
Procedure  
To enter the TDS conversion factor follow the instructions in in Table 8-5.  
Table 8-5 Procedure for Entering TDS Conversion Factor  
Step  
1
Screen  
Action  
MENU  
CONFIGURATION  
CALIBRATION  
Go to CALIBRATION | TDS CVSN FACTOR C1.  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
|
|
CALIBRATION  
If you plan to calibrate, do it before entering the  
theoretically derived TDS conversion factor (see previous  
page).  
CELL CAL FACTOR  
TDS CVSN FACTOR C1  
TDS CVSN FACTOR C2  
REMOVE CAL TRIM C1  
REMOVE CAL TRIM C2  
DIAGNOSTICS  
|
|
TDS CONVRSN FACTOR  
C1 0.5  
2
3
Enter the appropriate TDS conversion factor for the  
selected cell.  
DIAGNOSTICS  
| <-- | -->  
TDS CONVRSN FACTOR  
ACCEPTED AND SAVED  
If the factor entered is within the permitted range for the  
cell constant (see Table 5-10), the Analyzer/Controller will  
use the entered value.  
DIAGNOSTICS  
|
If the value is out-of-range, the Analyzer/Controller will  
force the value to the appropriate range limit.  
|
CALIBRATION  
4
Once an acceptable factor has been entered for Cell 1,  
repeat the process for Cell 2.  
CELL CAL FACTOR  
TDS CVSN FACTOR C1  
TDS CVSN FACTOR C2  
REMOVE CAL TRIM C1  
REMOVE CAL TRIM C2  
DIAGNOSTICS  
|
|
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
8-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
8-10  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics and Messages  
9. Diagnostics and Messages  
9.1 Overview  
Introduction  
This section contains information on status and alarm messages, as well as on diagnostics and  
system error messages. All these messages are displayed on the “alarm stripe” in “reverse  
video”, with the background lit and the letters black. If more than one message is active, the  
display will cycle through all the messages, then repeat.  
What’s in this section?  
The topics in this section are listed below.  
Topic  
See Page  
9-1  
9.1 Overview  
9.2 System Status Messages  
9.3 Process Alarm Messages  
9.4 On-Line Diagnostics and System Error Messages  
9.5 Off-Line Diagnostics  
9-2  
9-3  
9-4  
9-5  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
9-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
9.2 System Status Messages  
Related to operation of Analyzer/Controller  
System status messages, displayed on the alarm stripe, are related to normal operation of the  
Analyzer/Controller. These messages are listed in Table 9-1.  
Table 9-1 System Status Message  
MESSAGE  
AUTO CLEAN  
WHEN DISPLAYED  
This message is displayed temporarily and indicates that  
an AutoClean operation is active. All alarms and outputs  
are held for the duration of the AutoClean operation.  
To see how much time is remaining until the next  
scheduled AutoClean operation, use the DISPLAY key to  
access the special Time Remaining screen.  
This message indicates that a calibration operation has  
been unsuccessful.  
CAL ADJ FAILED  
This message is displayed during I/O setup, configuration,  
and calibration while the unit retrieves data from its  
memory.  
RETRIEVING DATA  
This message is displayed during I/O setup, configuration,  
and calibration while the unit processes data entered by  
the user.  
ENTERING DATA  
User input was either greater than or less than the  
allowable range.  
INPUT RANGE TOO HIGH  
INPUT RANGE TOO LOW  
9-2  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics and Messages  
9.3 Process Alarm Messages  
Alarms dependent on configuration  
Alarms are configurable as described in Section 5. When an alarm is active, it is always  
displayed on the alarm stripe. All available alarm messages are shown in Table 9-2.  
Table 9-2 Process Alarm Messages  
MESSAGE  
CELL 1 HIGH ALM  
CELL 1 LOW ALM  
CELL 2 HIGH ALM  
WHEN DISPLAYED  
Indicates alarm state, if the Analyzer/Controller is  
configured to alarm on the Cell 1 and/or Cell 2 value  
exceeding (high alarm) or falling below (low alarm) the  
configured alarm setpoint.  
CELL 2 LOW ALM  
CELL1 OR 2 HIGH ALM  
CELL 1 OR 2 LOW ALM  
Indicates alarm state, if the Analyzer/Controller is  
configured to alarm on a computed value exceeding (high  
alarm) or falling below (low alarm) the configured alarm  
setpoint.  
DIFFERENCE HIGH ALM  
DIFFERENCE LOW ALM  
% PASSAGE HIGH ALM  
% PASSAGE LOW ALM  
% REJECT HIGH ALM  
% REJECT LOW ALM  
PPM CO2 HIGH ALM  
PPM CO2 LOW ALM  
RATIO HIGH ALM  
RATIO LOW ALM  
Indicates alarm state, if the Analyzer/Controller is  
configured to alarm on the temperature from Cell 1 and/or  
Cell 2 exceeding (high alarm) or falling below (low alarm)  
the configured alarm setpoint.  
TEMP 1 HIGH ALM  
TEMP 1 LOW ALM  
TEMP 2 HIGH ALM  
TEMP 2 HIGH ALM  
TEMP 1 OR 2 HIGH ALM  
TEMP 1 OR 2 LOW ALM  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
9-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
9.4 On-Line Diagnostics and System Error Messages  
Self-tests at 0.5 second intervals  
On-line diagnostics are self-tests which are performed automatically as a background task every  
1/2 second. The system error messages (displayed on the alarm stripe if a test is failed) are listed  
in Table 9-3.  
Output affected by error  
In an error condition, the Analyzer will continue to perform all the functions it is capable of  
except that the output signal(s) will go to 0%. (Output signal manipulation does not occur with  
calibration failure or clock reset diagnostics.)  
Table 9-3 On-Line Diagnostics and System Error Messages  
SYSTEM ERROR  
MESSAGE  
WHEN  
TESTED  
CONDITION  
CORRECTIVE  
ACTION  
background  
background  
faulty analog-to-digital converter  
Have instrument serviced.  
A/D CONV FAILURE  
CELL 1 OUT OF RANGE  
or  
Cell reading is out of range for  
either the high or low limit of the  
selected cell constant. The  
cell’s display will stop updating  
until the condition and alarm  
message clears. Instead, to  
indicate an invalid display, the  
display will alternately flash 0.0  
and either the last known value  
or the high end limit for the cell  
constant.  
Check cell to make sure it is  
connected to terminals. Make  
sure cell constant is correct for  
expected conductivity range.  
CELL 2 OUT OF RANGE  
background  
Cell temperature compensator  
faulty  
Check cell (see Note 1).  
CELL 1 TEMP ERROR  
or  
CELL 2 TEMP ERROR  
CLOCK RESET  
after power  
failure  
Clock is on, but time may be  
incorrect, or  
Reset the time,  
or  
clock is off as a result of the  
power failure.  
if clock is off, turn it on and reset  
the time.  
at reset  
faulty EPROM  
Have instrument serviced.  
EPROM ERROR  
background  
data in non-volatile memory has  
been changed by a fault  
Power down, check power supply  
and wiring for high voltage  
spikes. Have instrument  
serviced, if necessary.  
RAM DATA CORRUPTED  
at reset  
at reset  
faulty RAM  
Have instrument serviced.  
Have instrument serviced.  
RAM ERROR  
faulty RAM in microprocessor  
RAM ERROR IN CPU  
Note 1: Temperature compensator should be 8550 ohms at 25 °C. If temperature compensator  
fails, a conductivity instrument will be forced to its display range high limit. A resistivity  
instrument will be forced to the low end of the range or zero.  
9-4  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics and Messages  
9.5 Off-Line Diagnostics  
Introduction  
The following off-line diagnostics can be initiated by the operator:  
display test (see 9.5.1)  
keyboard test (see 9.5.1)  
output tests (see 9.5.2)  
9.5.1 Display and Keyboard Test  
Procedure  
The procedure for initiating the display and keyboard test is described in Table 9-4.  
Table 9-4 Procedure For Initiating Display and Keyboard Tests  
Step  
1
Screen  
Action  
MENU  
Go to MAINTENANCE | OFF-LINE FUNCTIONS |  
DISPLAY TEST.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
MAINTENANCE  
OFFLINE FUNCTIONS  
INSTRUMENT SETUP  
LEAD RESISTANCE COMP  
DIAGNOSTICS  
| PREV | NEXT  
OFF-LINE FUNCTIONS  
DISPLAY TEST  
KEYBOARD TEST  
OUTPUT/RELAY TEST  
OUTPUT CALIBRATON  
DIAGNOSTICS  
| PREV | NEXT  
DISPLAY TEST  
2
Pressing the specified function key will display a test  
pattern on the first line of the display. Press the key  
repeatedly to test the rest of the display.  
DIAGNOSTICS  
|
After confirming that each line of the display is  
functional, press the MENU key to return to the off-line  
functions menu.  
DTEST|  
DISPLAY TEST  
F1=TOGGLE MENU=ABORT  
DIAGNOSTICS  
|
|
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
9-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Step  
3
Screen  
Action  
OFF-LINE FUNCTIONS  
DISPLAY TEST  
Go to MAINTENANCE | OFF-LINE FUNCTIONS |  
KEYBOARD TEST.  
KEYBOARD TEST  
OUTPUT/RELAY TEST  
OUTPUT CALIBRATON  
DIAGNOSTICS  
| PREV | NEXT  
KEYBOARD TEST  
4
Press any key (except MENU) to display the name of  
PRESS A KEY  
TO TEST  
the key.  
Use MENU to exit.  
MENU TO EXIT  
DIAGNOSTICS  
|
|
9-6  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics and Messages  
9.5.2 Output Tests  
Procedure  
Use the procedure in Table 9-5 to test the relays and analog outputs.  
Testing the analog outputs requires ammeters or voltmeters as appropriate.  
WARNING  
While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not open  
the case while the unit is powered.  
Do not access the output terminal as described below while the unit is powered. More than one  
disconnect switch may be required to remove power.  
CAUTION  
Consider the effects on your process before testing relays with controlled devices active.  
Table 9-5 Procedure for Testing Relays and Analog Outputs  
Step  
1
Screen  
Action  
Turn off the power to the Analyzer/Controller. More than  
one switch may be required to remove the power.  
2
With the power off open the case:  
Grasp the bottom center portion of the front bezel and pull it  
downward and toward you slightly to disengage the bottom of  
the bezel from the edge of the case.  
Lift the bezel gently to disengage it from the top edge of the  
case.  
Swing the bezel to the left. (The bezel and display assembly  
is mounted on pivot arms.) The safety cover with its warning  
label will be visible.  
3
4
5
With the power off remove the safety cover by removing the  
single screw holding it in place.  
Label and remove the field wiring connected to the analog  
output terminals.  
Feed the test wiring through the conduit hole in the case and  
connect the appropriate type of meter to each of the analog  
output terminals to be tested. Be sure to observe correct  
polarity.  
6
Close the case and power up the unit. Do not connect power  
until the case is closed.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
9-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Step  
7
Screen  
Action  
MENU  
Go to MAINTENANCE | OFF-LINE FUNCTIONS |  
OUTPUT/RELAY TEST.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
MAINTENANCE  
OFFLINE FUNCTIONS  
INSTRUMENT SETUP  
LEAD RESISTANCE COMP  
DIAGNOSTICS  
| PREV | NEXT  
OFF-LINE FUNCTIONS  
DISPLAY TEST  
KEYBOARD TEST  
OUTPUT/RELAY TEST  
OUTPUT CALIBRATON  
DIAGNOSTICS  
| PREV | NEXT  
OUTPUT TEST  
Select the combination of output values to be supplied to the  
terminals, or go to the next window for additional choices.  
01 02 03 R1R2R3R4  
0 0 0 A A A A  
25 0 0 A A A A  
50 25 0 A A A A  
75 50 25 A A N A  
DIAGNOSTICS  
When one of the rows of output combinations has been  
selected, observe the meter readings to check that the analog  
output values match those selected. Also listen for the “click”  
of an energized relay.  
|
| NEXT  
If the results of the analog output test are not satisfactory,  
recalibrate the output’s zero and span using the procedure in  
10.2.  
If the results of the relay output test are not satisfactory,  
consult Honeywell for assistance.  
OUTPUT TEST  
01 02 03 R1R2R3R4  
100 75 50 A A A N  
0 100 75 A N A A  
0 0 100 A A N A  
0 0 0 A N N N  
DIAGNOSTICS  
9
Select one of these combinations, or use the specified function  
key to return to the previous display.  
| PREV |  
10  
11  
If only one meter is available, so that only one output can be  
tested at a time, then for each analog output repeat Steps 1  
through 9, including powering down the unit before  
changing the connections to the output terminals.  
When output testing has been completed, re-install field wiring  
removed in Step 4. Disconnect power before opening case.  
12  
13  
Replace the safety cover and secure it with the screw.  
Close the case and power up the unit. Do not apply power  
until case is closed.  
9-8  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
10. Maintenance  
10.1 Overview  
Introduction  
This section provides instructions for a variety of maintenance tasks.  
What’s in this section?  
The topics in this section are listed below.  
Topic  
See Page  
10-1  
10.1 Overview  
10.2 Output Calibration  
10-3  
10.3 Specifying a Tagname or Other Display String  
10.4 Viewing Product Information and Changing Model Number Stored in Memory  
10.5 Adjusting the Screen Contrast  
10-6  
10-7  
10-9  
10.6 Entering a Password for Security  
10-10  
10-11  
10-12  
10-14  
10.7 Resetting All Configuration and Calibration Values to Factory Settings  
10.8 Entering Values for Lead Resistance Compensation  
10.9 Replacing the Fuse  
WARNING  
While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not open  
the case while the unit is powered. More than one disconnect switch may be required to remove  
power.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
10-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
ATTENTION  
Most of the operations described in this section are performed with the unit off-line (that is, not  
connected to the process). However, some of the operations described in this section are  
performed with the Analyzer/Controller on-line. Your process may be affected as soon as you  
initiate the maintenance function.  
ATTENTION  
This equipment contains devices that can be damaged by electrostatic discharge (ESD). The  
damage incurred may not cause the device to fail completely, but may cause early failure.  
Therefore, it is imperative that assemblies containing static sensitive devices be carried in non-  
conductive plastic bags. When adjusting or performing any work on such assemblies, grounded  
work stations and wrist straps must be used. If soldering irons are used, they must also be  
grounded.  
A grounded work station is any conductive or metallic surface connected to an earth ground,  
such as a water pipe, with a 1/2 to 1 megohm resistor in series with the ground connection. The  
purpose of the resistor is to current limit an electrostatic discharge and to prevent any shock  
hazard to the operator. The steps indicated above must be followed to prevent damage and/or  
degradation, which may be induced by ESD, to static sensitive devices.  
10-2  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
10.2 Output Calibration  
Introduction  
The 9782 is available with one or more optional analog outputs. The output signals can be  
adjusted to trim the high and low output current or voltage values over a range of ±0.4% of span  
to compensate for component tolerance variations.  
Required equipment  
Output calibration involves connecting a meter to the Analyzer/Controller’s output terminals.  
The meter required for output calibration depends on the type of outputs.  
current outputs: current meter capable of resolving 0.01 mA over the range 0 to 20 mA dc  
voltage outputs: a 250 ohm ±0.05% shunt and a volt meter (capable of measuring 1 to 5 Vdc  
within 1 mV)  
A screwdriver to fit the terminal block screws is also required.  
Procedure  
To calibrate outputs, follow the procedure described in Table 10-1. The output terminals are  
inside the case as shown in Figure 6-1.  
WARNING  
While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not open  
the case while the unit is powered. Do not access the output terminal as described below while  
the unit is powered.  
WARNING  
A disconnect switch must be installed to break all current carrying conductors. Turn off power  
before working on conductors. Failure to observe this precaution may result in serious personal  
injury.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
10-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Table 10-1 Procedure for Calibrating Outputs  
Step  
1
Screen  
Action  
Turn off the power to the Analyzer/Controller. More than  
one switch may be required to disconnect power.  
2
With the power off open the case:  
Grasp the bottom center portion of the front bezel and pull it  
downward and toward you slightly to disengage the bottom of  
the bezel from the edge of the case.  
Lift the bezel gently to disengage it from the top edge of the  
case.  
Swing the bezel to the left. (The bezel and display assembly is  
mounted on pivot arms.) The safety cover with its warning label  
will be visible.  
3
4
5
With the power off remove the safety shield by removing the  
single screw holding it in place.  
Label and remove the field wiring connected to the analog  
output terminals.  
Feed the test wiring through conduit hole in the case and  
connect the appropriate type of meter to the terminals for the  
output to be calibrated. Be sure to observe correct polarity.  
6
7
Close the case and power up the unit.  
MENU  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
Go to MAINTENANCE | OFF-LINE FUNCTIONS | OUTPUT  
CALIBRATION.  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
MAINTENANCE  
OFFLINE FUNCTIONS  
INSTRUMENT SETUP  
LEAD RESISTANCE COMP  
DIAGNOSTICS  
| PREV | NEXT  
OFF-LINE FUNCTIONS  
DISPLAY TEST  
KEYBOARD TEST  
OUTPUT/RELAY TEST  
OUTPUT CALIBRATION  
DIAGNOSTICS  
|
|
10-4  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Step  
8
Screen  
Action  
OUTPUT1 SPAN  
The first display will reflect the quantity of analog outputs  
supported by the hardware.  
OUTPUT1 ZERO  
OUTPUT2 SPAN  
OUTPUT2 ZERO  
OUTPUT3 SPAN  
OUTPUT3 ZERO  
DIAGNOSTICS  
Select the value to be calibrated.  
Go to the next screen.  
|
|
SPAN CAL OUTPUT1  
9
Use the UP and DOWN keys to correct the reading on the test  
meter.  
USE UP/DOWN ARROW  
KEYS TO ADJUST  
ENTER TO SAVE  
To save the correction press ENTER.  
DIAGNOSTICS  
|
|
OUTPUT CALIBRATION  
SAVED  
10  
This screen confirms that the value was saved.  
MENU TO EXIT  
DIAGNOSTICS  
|
|
11  
12  
To calibrate additional values, repeat Steps 1 through 10,  
including powering down the unit before changing the  
connections to the output terminals.  
When output calibration has been completed, re-install field  
wiring removed in Step 4. Disconnect power before opening  
case.  
13  
14  
Replace the safety cover and secure it with the screw.  
Close the case and power up the unit. Do not apply power  
until the case is closed.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
10-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
10.3 Specifying a Tagname or Other Display String  
Introduction  
The real-time displays of process values show the instrument’s tagname (or other configurable  
fixed nineteen-character string) at the top of the screen.  
Procedure  
Use the procedure in Table 10-2 to specify the text to be displayed on the alarm stripe when no  
alarms or other messages are active.  
Table 10-2 Procedure for Specifying a Tagname or Other Display String  
Step  
1
Screen  
Action  
MENU  
Go to MAINTENANCE | INSTRUMENT SETUP |  
INSTRUMENT TAGGING.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
MAINTENANCE  
OFFLINE FUNCTIONS  
INSTRUMENT SETUP  
LEAD RESISTANCE COMP  
DIAGNOSTICS  
| PREV | NEXT  
INSTRUMENT SETUP  
INSTRUMENT TAGGING  
PRODUCT INFO  
SCREEN CONTRAST  
SECURITY  
RESET UNIT  
DIAGNOSTICS  
| PREV | NEXT  
UPDATE TAG DATA  
HONEYWELL INC  
2
The text string currently stored in memory will be  
displayed.  
If desired, edit the string (nineteen characters maximum)  
using the UP and DOWN keys, and the <-- and --> function  
keys.  
DIAGNOSTICS  
| <-- | -->  
Press ENTER or MENU to exit.  
10-6  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
10.4 Viewing Product Information and Changing Model Number  
Stored in Memory  
Introduction  
During I/O setup (see Section 5) the software will display prompts only for those features the  
hardware can support. To do this, the software uses the model number of the unit stored in  
memory at the factory. Each combination of features has a unique model number (see Section 2).  
Therefore, if you add an option card to the 9782, it is very important that you change the model  
number stored in memory using the procedure described below.  
ATTENTION  
A special password is required to change the stored model number. If the password was not  
supplied with the new option card, call the Honeywell Technical Assistance Center.  
Before calling for technical support, follow the procedure in Table 10-4 to determine the version  
of the software in the Analyzer/Controller. The support team will ask for this information.  
Procedure  
Use the procedure in Table 10-4 to see the software version and model number, and to change  
the model number if required.  
Table 10-3 Procedure for Viewing Product Information and  
Changing the Stored Model Number  
Step  
1
Screen  
Action  
MENU  
Go to MAINTENANCE | INSTRUMENT SETUP |  
PRODUCT INFO.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
MAINTENANCE  
OFFLINE FUNCTIONS  
INSTRUMENT SETUP  
LEAD RESISTANCE COMP  
DIAGNOSTICS  
| PREV | NEXT  
INSTRUMENT SETUP  
INSTURMENT TAGGING  
PRODUCT INFO  
SCREEN CONTRAST  
SECURITY  
RESET UNIT  
DIAGNOSTICS  
|
|
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
10-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Step  
2
Screen  
Action  
PRODUCT INFORMATION  
The software version is read-only information.  
SOFTWARE VER: A1.0  
MODEL NUMBER:  
9782X-nn-Xn-XXXX  
If the model number displayed no longer matches the  
hardware because you have added or changed an option  
card, use the specified function key to edit the model  
number. (See Section 2 for available model number  
selections.)  
DIAGNOSTICS  
|
EDIT |  
FACTORY PASSWORD  
3
4
Note that a special password is required to change the  
model number. (This is not the same as the security  
password described in 10.7.) Enter it when prompted.  
OOOO  
ENTER PASSWORD  
This password is included with the instructions for installing  
the new card.1  
DIAGNOSTICS  
|
|
MODEL NUMBER:  
9782X-nn-Xn-XXXX  
DIAGNOSTICS  
ATTENTION: Changing the model number stored in  
memory will do a “cold reset” of the unit, returning ALL  
configuration values to the factory settings shown in Table  
5-1. During a cold reset some I/O setup values will also  
be returned to the defaults, and the cal factor will be reset  
to match the cell constant. The calibration trim or TDS  
conversion factor will be removed.  
|
|
The “X” values in the model number are “don’t care”  
conditions, because these characters do not affect I/O  
setup and configuration parameters.  
Once you have completed the model number change,  
press ENTER to accept the change.  
To cancel the edit and avoid resetting the unit, press the  
MENU key.  
1
If you do not have the factory password, call the Technical Assistance Center at 1-800-423-9883.  
10-8  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
10.5 Adjusting the Screen Contrast  
Introduction  
The display contrast can be adjusted to suit the viewing conditions at the installation site.  
Procedure  
To adjust the display contrast, follow the procedure in Table 10-4.  
Table 10-4 Procedure for Adjusting the Screen Contrast  
Step  
1
Screen  
Action  
MENU  
Go to MAINTENANCE | INSTRUMENT SETUP | SCREEN  
CONTRAST.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
MAINTENANCE  
OFFLINE FUNCTIONS  
INSTRUMENT SETUP  
LEAD RESISTANCE COMP  
DIAGNOSTICS  
| PREV | NEXT  
INSTRUMENT SETUP  
INSTURMENT TAGGING  
PRODUCT INFO  
SCREEN CONTRAST  
SECURITY  
RESET UNIT  
DIAGNOSTICS  
|
|
SCREEN CONTRAST  
20  
2
Use the UP and DOWN keys to adjust the contrast  
between 0 and 20.  
USE UP/DOWN ARROW  
KEYS TO ADJUST  
ENTER TO SAVE  
When the displayed contrast is acceptable, use the  
ENTER key to return to the instrument setup menu. The  
new contrast setting will be saved automatically.  
DIAGNOSTICS  
|
|
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
10-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
10.6 Entering a Password for Security  
Introduction  
I/O setup, configuration, calibration and maintenance functions can be password-protected. The  
password can be any number between 1 and 9999. (When the password is zero, the operator will  
not be prompted to enter a password.)  
ATTENTION  
The configurable password entered here is not the same as the special password needed to edit  
the model number stored in memory as described in 10.5.  
Procedure  
Follow the procedure in Table 10-5 to enter a password.  
Table 10-5 Procedure for Entering a Password  
Step  
1
Screen  
Action  
MENU  
Go to MAINTENANCE | INSTRUMENT SETUP |  
SECURITY.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
MAINTENANCE  
OFFLINE FUNCTIONS  
INSTRUMENT SETUP  
LEAD RESISTANCE COMP  
DIAGNOSTICS  
| PREV | NEXT  
INSTRUMENT SETUP  
INSTURMENT TAGGING  
PRODUCT INFO  
SCREEN CONTRAST  
SECURITY  
RESET UNIT  
DIAGNOSTICS  
|
|
PASSWORD SECURITY  
2
To enable the security function, select YES.  
ENABLE SECURITY: YES  
PASSWORD:  
0000  
The current password will be displayed. It can be edited  
using this screen.  
DIAGNOSTICS  
|
|
10-10  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
10.7 Resetting All Configuration and Calibration Values to Factory  
Settings  
Introduction  
A “cold reset” function is available to return ALL configuration values to the factory settings  
shown in Table 5-1. During a cold reset the cal factor will be changed to match the cell constant.  
The calibration trim or TDS conversion factor will be removed. All alarms will be cleared and  
outputs will go to zero. Some I/O setup values will also be returned to the factory defaults.  
This “cold reset” contrasts with the “warm reset” that occurs following loss of power and at the  
end of system setup (see Section 5). During a warm reset, alarms are not cleared, and outputs are  
maintained at their last value prior to the loss of power or initiation of the cold reset. A warm  
reset has no effect on configuration or calibration values.  
Procedure  
To do a cold reset, returning all configuration and calibration values to their factory settings,  
follow the procedure in Table 10-6.  
Table 10-6 Procedure for Resetting All Configuration to Factory Settings  
Step  
1
Screen  
Action  
MENU  
Go to MAINTENANCE | INSTRUMENT SETUP | RESET  
UNIT.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
MAINTENANCE  
OFFLINE FUNCTIONS  
INSTRUMENT SETUP  
LEAD RESISTANCE COMP  
DIAGNOSTICS  
| PREV | NEXT  
INSTRUMENT SETUP  
INSTURMENT TAGGING  
PRODUCT INFO  
SCREEN CONTRAST  
SECURITY  
RESET UNIT  
DIAGNOSTICS  
|
|
UNIT RESET  
ATTENTION  
ALL PARAMETERS  
WILL BE RESET TO  
FACTORY DEFAULTS!  
2
Pressing the specified function key at this point will  
perform a cold reset as described above.  
As the reset is completed. the display will go blank, then  
for a few seconds product information will be displayed.  
Finally, the display will return to showing process values.  
DIAGNOSTICS  
|
RESET|  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
10-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
10.8 Entering Values for Lead Resistance Compensation (Wide Range  
Only)  
Introduction  
If you use standard Honeywell cell lead lengths of 7 or 20 feet connected directly to the  
Analyzer/Controller, no compensation for lead resistance is necessary. Similarly, if a junction  
box is used to extend the leads up to 150 feet, no compensation is required. However, if longer  
leads are used (greater than 150 feet), signal quality can be adversely affected unless you enter  
information that will permit the 9782 to compensate for lead resistance.  
If you use a single wire gauge (12, 14, 16, or 18 AWG) in a length up to 1500 feet, simply  
specify the gauge and length as described in Table 10-9.  
If mixed wired gauges are used, or lead length or wire gauge are not within the stated ranges, the  
9782 can still perform the compensation. However, you must first calculate the lead resistance,  
then put it in terms of the available settings for AWG gauge and length.  
The resistance of each available gauge choice (in copper wire) is:  
12 AWG = 1.6 ohms per 1000 feet  
14 AWG = 2.5 ohms per 1000 feet  
16 AWG = 4.0 ohms per 1000 feet  
18 AWG = 6.4 ohms per 1000 feet  
For example, suppose each lead between the cell and Analyzer/Controller consists of 500 feet of  
12 gauge wire and 1000 feet of 18 gauge wire.  
1000 ft of 18 AWG  
500 ft. of 12 AWG  
9782  
Honeywell  
Junction  
Box  
Conductivity Cell  
Figure 10-1 Example of a Conductivity Loop  
Because there are two different types of wire used in each lead to the cell in this example, the  
total lead resistance is calculated as follows:  
(2 x 0.5 x 1.6) + (2 x 1 x 6.4) = 14.4 ohms  
Since the 9782 only allows entry of one wire gauge type, we allow for the worst-case condition  
by dividing the total resistance by the resistance per thousand feet of the higher resistance gauge  
wire. In our example this would be:  
14.4 ohms ÷ 6.4 ohms per thousand feet of 18 AWG wire = 2,250 feet  
The length to enter is one-half this number, 1125 feet, because the 9782 Analyzer/Controller  
already accounts for the fact that there is always a pair of conductor wires in the system loop.  
Therefore, in our example we would use the procedure in Table 10-9, and specify the wire gauge  
as 18 AWG and the length as 1125 feet.  
10-12  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Procedure  
To use the 9782’s automatic compensation for lead resistance, enter the necessary values using  
the procedure in Table 10-7.  
Table 10-7 Procedure for Entering Values for Lead Resistance Compensation  
Step  
1
Screen  
Action  
MENU  
Go to MAINTENANCE | LEAD RESISTANCE COMP.  
CONFIGURATION  
CALIBRATION  
MAINTENANCE  
I/O SETUP  
DIAGNOSTICS  
| PREV | NEXT  
MAINTENANCE  
OFFLINE FUNCTIONS  
INSTRUMENT SETUP  
LEAD RESISTANCE COMP  
DIAGNOSTICS  
| PREV | NEXT  
LEAD RESISTANCE COMP  
2
Use this screen to specify the gauge used for leads to both  
cells, and the length of one lead to each cell.  
WIRE GAUGE  
WIRE LENGTH  
CELL1  
12AWG  
100FT  
150FT  
If mixed gauges are used, or the length exceeds 1999 feet,  
refer to the Introduction of this sub-section for guidance in  
computing the values to enter here.  
CELL2  
DIAGNOSTICS  
| PREV | NEXT  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
10-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
10.9 Replacing the Fuse  
Introduction  
Figure 10-1 shows the location of the power fuse on the power supply card in the right-most  
position inside the case. (See Section 11 for the part number of the kit containing fuses.)  
3
4
5
6
7
8
9
Figure 10-2 Location of Power Fuse  
WARNING  
Disconnect power before opening the instrument case to replace the fuse. A potentially lethal  
shock hazard exists inside the case if the unit is opened while powered. More than one switch  
may be required to disconnect power.  
ATTENTION  
To avoid potentially serious damage to expensive components, a blown fuse should not be  
replaced until the cause has been determined by a qualified service technician.  
Procedure  
To replace the fuse follow the procedure in Table 10-8.  
10-14  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Table 10-8 Procedure for Replacing the Fuse  
Action  
Step  
1
2
Determine why the fuse blew and correct the problem.  
Check the fuse’s packaging (in miscellaneous parts kit 51198177-501) to be sure that the  
fuse is correct for the voltage at your site.  
3
4
Turn off the power to the Analyzer/Controller. More than one switch may be required to  
remove power.  
With the power off open the case:  
Grasp the bottom center portion of the front bezel and pull it downward and toward you  
slightly to disengage the bottom of the bezel from the edge of the case.  
Lift the bezel gently to disengage it from the top edge of the case.  
Swing the bezel to the left. (The bezel and display assembly is mounted on pivot arms.)  
The safety cover with its warning label will be visible.  
5
6
With the power off remove the safety cover by removing the single screw holding it in  
place.  
If more room is needed, remove the circuit card retainer bracket at the top of the case and  
pull the circuit card assembly partially out of the instrument case. (It may be necessary to  
temporarily disconnect leads from some terminals to do so.)  
7
8
Remove the old fuse and insert the new one.  
If any leads were disconnected in Step 6, reconnect them.  
Replace the safety cover and secure it with the screw.  
9
10  
Close the case and power up the unit. Do not apply power until the case is closed.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
10-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
10-16  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Accessories and Replacement Parts List  
11. Accessories and Replacement Parts List  
11.1 Overview  
Introduction  
This section provides part numbers for field-replaceable parts and for accessories.  
What’s in this section?  
The topics in this section are listed below.  
Topic  
See Page  
11-1  
11.1 Overview  
11.2 Part Numbers  
11-2  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
11-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
11.2 Part Numbers  
Introduction  
Part numbers for field-replaceable parts and accessories are provided in Table 11-1.  
Table 11-1 Part Numbers  
Kit/Part Number  
51198177-501  
Description  
Parts Included in Kit  
Quantity  
PARTS COMMON TO ALL 9782 ANALYZERS  
Miscellaneous parts for all 9782  
controller/analyzers  
120 volt fuse  
240 volt fuse  
case gasket  
1
1
1
1
1
2
1
1
1
1
upper card retainer bracket  
lower card retainer bracket  
screws for retainer brackets  
card  
installation instructions  
card  
Output card: single current or voltage  
output  
Output card: single current or voltage  
output with two additional hermetically  
sealed relays  
51198063-501  
51198064-501  
installation instructions  
Output card: two additional general  
purpose relays and no analog outputs  
For replacement of existing card  
(2 current outputs) on pH/cond. analyzer  
For replacement of existing card  
(3 current outputs) on pH/cond. analyzer  
Non Backlit LCD Display  
card  
installation instructions  
card  
installation instructions  
card  
installation instructions  
display  
installation instructions  
display  
installation instructions  
bezel  
installation instructions  
card  
installation instructions  
card  
installation instructions  
card  
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
4
1
1
1
1
51198065-501  
51198066-501  
51198067-501  
51198176-501  
51198176-502  
51205783-501  
51205501-501  
51205502-501  
51205503-501  
Backlit LCD Display w/solid state LED  
Bezel (front cover w/keypad)  
Power Supply PC  
Microprocessor PC  
Backplane PC  
installation instructions  
cable  
Display cable  
Case and gasket  
51198095-501  
51198068-501  
case (no bezel)  
case gasket  
filter cores  
Power line filter kit  
079163  
084711  
installation instructions  
2 U-bolts for 1 in. pipe  
2 U-bolts for 2 in. pipe  
mounting plate  
hardware  
Pipe/wall mounting kit  
Panel mounting kit  
Reducer panel kit  
panel bracket  
fuse  
hardware  
1
056348  
056349  
installation instructions  
panel  
1
1
11-2  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Accessories and Replacement Parts List  
Kit/Part Number  
51198066-503  
Description  
Parts Included in Kit  
Quantity  
PARTS USED BY CONDUCTIVITY 9782 ANALYZER ONLY  
For upgrade of conductivity analyzer  
Wide Range: 2 current outputs  
card  
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
prom  
installation instructions  
card  
For upgrade of conductivity analyzer  
Standard Range: 2 current outputs  
51198066-504  
51198067-503  
51198067-504  
prom  
installation instructions  
card  
For upgrade of conductivity analyzer  
Wide Range: 3 current outputs  
prom  
installation instructions  
card  
For upgrade of conductivity analyzer  
Standard Range: 3 current outputs  
prom  
installation instructions  
card  
installation instructions  
card  
installation instructions  
assembly  
installation instructions  
assembly  
Replacement conductivity standard range  
input card  
Replacement conductivity wide range  
input card  
Bezel ass’y. for Non Backlit Display Units  
Wide & Standard Range Conductivity  
Bezel ass’y. for Backlit Display Units  
Wide & Standard Range Conductivity  
51198074-501  
51198087-501  
51198174-501  
51205779-501  
installation instructions  
AVAILABLE CABLES  
Part Number  
Description  
Parts Included  
Note  
Extension lead wires up to 500 ft - standard range models  
834059  
3-conductor, 18 gauge cable (Belden  
9493)  
cable  
cable  
Note 1  
Note 1  
835024  
coaxial cable (Belden 9259)  
Extension lead wires up to 1000 ft - standard range models  
834055  
835024  
4-conductor (3 used), 16 gauge cable  
coaxial cable (Belden 9259)  
cable  
cable  
Note 1  
Note 1  
Extension lead wires up to 500 ft - wide range models  
31834052 4-conductor, 18 gauge cable  
Extension lead wires up to 1000 ft - wide range models  
cable  
Note 1  
834055  
Part Number  
31316260  
4-conductor, 16 gauge cable  
cable  
Note 1  
Description  
Parts Included  
Quantity  
aluminum junction box for use with  
additional cable for distances beyond  
150 ft  
junction box  
installation instructions  
1
1
Note 1: Order cable by the length in whole foot increments.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
11-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
11-4  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Proportional Control Tuning  
Appendix A  
Proportional Control Tuning  
Adjust effective proportional band width  
The only adjustment in proportional-only control is the effective proportional band width. For  
startup, enter the desired control setpoints for discrete control types (or the value representing  
0% in CAT control). Set the proportional band limits (value for 100% output) to the extreme  
process values anticipated before treatment of the batch.  
Treatment takes too long, but no unacceptable overshoot  
After observing treatment performance, some improvement may be possible. If treatment takes  
too long, and there has been no unacceptable overshoot beyond the setpoint, make one or more  
the following changes:  
Increase the reagent delivery capacity.  
Increase the reagent concentration.  
Decrease the proportional band width by adjusting the proportional band limit closer to the  
setpoint.  
Unacceptable overshoot  
If unacceptable overshoot occurs, make one or more of the following changes:  
Increase mixing in the treatment tank.  
Reduce the reagent delivery capacity.  
Reduce the reagent concentration.  
Widen the proportional band by adjusting the proportional band limit further from the  
setpoint.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Appendix A - 1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Appendix A - 2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Two-Cell Applications  
Appendix B  
Two-Cell Applications  
Ion Exchange  
Ion exchange operations can achieve especially precise control using the conductivity ratio of  
two points with each bed. Ratio measurement accounts for feedwater variations when the  
upstream point is measured at the cation bed inlet. With the upstream point in the bed as shown  
for following stages, it can identify exhaustion before breakthrough.  
Inlet  
Cell 2  
Cell 2  
Cell 1  
Cell 2  
Cell 1  
Cell 1  
Cation Bed  
D I Water  
a/n 23188  
Anion Bed  
Mixed Bed  
Reverse Osmosis  
Reverse Osmosis efficiency is monitored by comparing inlet and outlet conductivity (or TDS).  
Automatic calculations of Percent Rejection or Percent Passage are provided. If readout is in  
resistivity, cell locations are interchanged. Temperature readout assists with normalized  
performance comparisons.  
C ell 1  
C ell 2  
R O Unit  
Feed  
Perm eate  
a/n 23189  
C oncentrate  
Cell1  
Cell2  
x100  
Percent Passage =  
Typical Range is 0 to 20%  
Typical Range is 80 to 100%  
Cell1  
Cell2  
(1−  
)x100  
Percent Rejection =  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Appendix B - 1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Conductivity/Resistivity/TDS Difference  
Conductivity/Resistivity/TDS difference using redundant cells on critical processes can provide a  
valuable diagnostic capability. If the difference in measurements exceeds the alarm points, an  
operator is summoned for corrective action. Monitoring may be switched to the alternate cell  
during maintenance. For deviation in either direction, two different alarms (+ and -) are used. A  
difference kind of diagnostic can be provided by a precision check resistor in place of one cell to  
give continuous Analyzer/Controller checking at one value. See also Appendix D.  
9782 Analyzer  
OutputSignal  
Difference Alarm  
Cell 2  
Cell 1  
Process  
a/n 23190  
Parts Rinsing  
Parts rinsing is usually controlled by conductivity to obtain adequate rinsing without wasting  
excess water, whether a single stage or a counter-current series of tanks. The two-cell ratio  
approach can determine whether inadequate rinsing is due to low flowrate or due to poor supply  
water quality.  
Cell 1  
Cell 2  
Parts Flow  
WasteWater  
a/n 23191  
Rinse  
Water  
Cell1  
Cell2  
Conductivity Ratio =  
Typical Ratio Range is 0.1 to 1.1  
Appendix B - 2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Two-Cell Applications  
Softener Monitoring  
Softener monitoring by conductivity ratio gives a continuous indication of performance. Sodium  
is typically more conductive than the hardness minerals it displaces, yielding a higher  
conductivity at the outlet. A ratio approaching 1 indicates that hardness ions are breaking through  
and that regeneration is needed.  
(Hard) Water Supply  
Cell 2  
Cell 1  
a/n 23192  
(Soft) Treated Water  
Cell1  
Cell2  
Softening Ratio =  
Typical Ratio Range is 1 to 1.25  
Steam Power Measurements  
The three conductivity measurements in power plants relate to water chemistry parameters as  
follows:  
Specific conductivity -- chemical treatment level  
Cation conductivity -- total anion contaminants  
Degassed conductivity -- non-volatile anion contaminants  
Cation minus degassed conductivities -- carbon dioxide  
The 9782 Standard Range Analyzer/Controller includes a complete menu of field-selectable  
functions for these determinations. In addition to specialized high purity temperature  
compensation for specific and cation conductivity, the 9782 includes reboiler measurement  
functions to automate ASTM Test Method D4519 for determining anions and carbon dioxide.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Appendix B - 3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Specific  
Conductivity  
Cation  
Conductivity  
9782  
Analyzer  
Sample  
Cell 1  
Cell 2  
Cation  
Exchanger  
Specific  
Degassed  
Conductivity  
(Anions)  
Conductivity  
Cation  
Conductivity  
7082  
Analyzer  
9782  
Analyzer  
Carbon Dioxide  
by Calculation  
Sample  
Cell 1  
Cation  
Exchanger  
Cell 2  
Reboiler  
a/n 23193  
Sodium Hydroxide & Hydrochloric Acid Concentration Measurements  
The measurement range of sodium hydroxide by conductivity is limited by temperature. The  
conductivity is limited by temperature. The conductivity of sodium hydroxide reaches a  
maximum value near 14% at 0º C and 29% at 100º C. Near the maximum there is poor resolution  
and no reliable way to know which side of the peak is being measured. Therefore, the 7082  
measurement range is restricted by temperature to assure reliable values. Maximum  
concentrations are 10% at 0º C, 13% at 25º C and 20% at 75º C and above, with linear  
interpolation between these points. Operation above these limits gives a flashing display.  
The measurement range of hydrochloric acid is restricted to less than 15.5% above 40º C and less  
than 18% below 40º C.  
Appendix B - 4 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Microprocessor Board Switch Settings  
Appendix C  
Microprocessor Board Switch Settings  
Introduction  
The information in this appendix will not be needed for normal set up and operation of the 9782  
Analyzer/Controller. It is included here for use during troubleshooting with the telephone  
assistance of the Honeywell Technical Assistance Center.  
Location of switches  
Two 8-position DIP switches are mounted on the microprocessor board (far left board). The  
switch locations are shown in Figure C-1.  
Switch 1  
Switch 2  
1
2
3
4
5
1
6
2
3
4
5
6
7
8
9
Figure C-1 Microprocessor Board DIP Switches  
Note that S1 switch (030480), CR1-CR8 and CR12-CR16 (188056) are not used as of June 1993.  
SW1 - all positions set to off  
Regardless of the options used, all eight switches in SW1 are set to off.  
SW2 - settings are dependent on options  
The settings of the eight switches in SW2 depend on the options used. The appropriate settings  
are shown in Table C-1.  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Appendix C - 1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Table C-1 Microprocessor Board Switch Settings  
Option  
SW2-1 SW2-2 SW2-3 SW2-4 SW2-5 SW2-6 SW2-7 SW2-8  
None  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
Single  
Parameter  
Current or  
Voltage  
Output  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
Auxiliary  
Relays  
OFF  
ON  
OFF  
OFF  
ON  
ON  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
OFF  
Two Analog  
Outputs  
OFF  
OFF  
Three  
ON  
Analog  
Outputs  
Appendix C - 2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using a Precision Check Resistor  
Appendix D  
Using a Precision Check Resistor  
Introduction  
The operation of the Analyzer/Controller can be verified by replacing the input from a cell with a  
precision check resistor across the Analyzer/Controller input terminals. In addition, an 8550  
ohm resistor (Honeywell Part No. 233300) can be wired in place of the inputs from the  
temperature compensator to simulate 25º C, the reference temperature. The unit will display a  
simulated “process value” appropriate for the check resistor installed. (Equations showing the  
relationship between resistor rating and displayed value are provided below.) If the displayed  
value is incorrect, the Analyzer/Controller should be serviced.  
This technique can be used two ways:  
Offline - Install the precision check resistor temporarily in place of the input from either cell  
to check the operation of the Analyzer/Controller. When correct operation has been verified,  
remove the resistor and replace the field wiring.  
Online - To provide a constant check of the Analyzer/Controller’s operation in a critical  
process, connect the conductivity cell to the Cell 1 input terminals; instead of a Cell 2 input,  
install a check resistor at the Cell 2 input terminals. The Cell 2 “process value” should  
always be the appropriate value for the resistor (see equations below). Configure an alarm to  
monitor this value.  
Set cal factor and calibration trim for ideal conditions  
When a check resistor is used instead of cell input, the Analyzer/Controller must be set for  
theoretically ideal conditions to achieve display of the appropriate value for the installed resistor.  
This means that you set the cell calibration factor to 1.00 and remove the calibration trim for the  
cell input being replaced by the check resistor.  
Calculations for conductivity, resistivity, and TDS  
To verify instrument operation at any point of measurement, calculate the check resistance  
needed to simulate that value. (It is assumed that you have selected a display measurement value  
that is within the range of your cell constant; see 2.1.5 for ranges.) The equation used depends  
on the measurement type. For concentration check values see the table on the following page.  
Conductivity check resistance (ohms) = Cell Constant (cm-1) x 106  
Conductivity (microSiemens/cm)  
Resistivity check resistance (ohms) = Cell Constant (cm-1) x Resistivity (ohm-cm)  
TDS check resistance (ohms) = Cell Constant (cm-1) x 106  
TDS (ppm)/TDS factor  
(TDS factor has units of ppm/microSiemens-cm-1)  
7/99  
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual Appendix D - 1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conductivity/Resistivity Analyzer/Controller  
Example 1: To determine the check resistor value needed to simulate conductivity measurement  
of 10 µS, use cell constant 0.1 and perform the following calculation:  
10 k ohms = (0.1) x (1,000,000)  
10  
Example 2: To determine the check resistor value needed to simulate resistivity measurement of  
10 M ohms, use cell constant 0.01 and perform the following calculation:  
100 K ohms = (0.01) x (10,000,000)  
Concentration values  
If a wide range model is used for concentration measurement, obtain the appropriate check  
resistance value from the table below.  
Data for Concentration Range Measurements  
Material/Weight % Concentration  
Simulation Resistance (ohms) @ 25º C  
Cell Constant  
10  
25  
50  
Hydrochloric Acid (HCl)  
0
1
4
242.5  
68.9  
485.0  
137.7  
Sulfuric Acid (H2SO4)  
0
1
4
215.5  
56.0  
538.7  
140.0  
1077.4  
280.0  
Sodium Chloride (NaCl)  
0
1
4
574.1  
195.2  
1435.1  
398.0  
2870.3  
796.1  
Sodium Hydroxide (NaOH)  
0
1
4
189.2  
54.0  
473.0  
135.1  
946.1  
270.1  
Appendix D - 2 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual  
7/99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Industrial Automation and Control  
Honeywell, Inc.  
1100 Virginia Drive  
Fort Washington, Pennsylvania 19034  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Haier Beverage Dispenser JC 82GB User Manual
Haier Washer HKS1000TXVE HKS1000TXME User Manual
Hamilton Beach Coffeemaker 47453 User Manual
Hans Grohe Outdoor Shower 16320XX1 User Manual
Hansol Multitech Computer Monitor D17CL User Manual
Honeywell Thermostat 50 9512 RL User Manual
HP Hewlett Packard Server DL180 G5 User Manual
Huawei Cordless Telephone ETS8121 User Manual
hussman Refrigerator P N 0420057_B User Manual
Ikelite Digital Camera IXUS65 User Manual