HP Hewlett Packard Network Router 2040D User Manual

Series 2040D Test Systems  
2040D  
Maintenance  
Manual  
Part Number #4200-0181  
Version 1.0  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CONTACT SPRING REPAIR ............................................................. 63  
USE OF PATCHBOARD SPRING REMOVAL TOOL (0000-2746)...... 64  
CONTACT SPRING REPLACEMENT (Conventional Method) . 65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Maintenance Overview  
System Overview  
Series 2040 Test System  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance Overview  
Series 2040 Maintenance Manual V2.00  
Series 2040 Test System  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Maintenance Overview  
SYSTEM MAINTENANCE OVERVIEW  
The Series 2040 Test System has been designed with extensive diagnostics and  
self testing capabilities. Each module and circuit board includes additional  
circuitry for calibration and self testing programs. In addition, a Selftest  
Assembly is provided to calibrate and functionally test every component in the  
test system. This assembly attaches to the fixture interface to provide the best  
accuracy and reliability for calibration and self-diagnostics.  
The Certification program uses a Calibration station containing a Digital  
Voltmeter (DVM) and Universal Counter/Timer, which are certified by the  
National Institute of Standards and Technology as a “Transfer Standard,” to  
calibrate the internal precision references of the tester. The Selftest Assembly  
uses a 18-bit D/A converter (TDAC) as an internal precision voltage standard.  
The internal precision time standard is a Temperature Compensated Crystal  
Oscillator located on the Time Measurement System board.  
The Calibration and Selftest routines calibrate the remainder of the system to  
the tester’s internal precision references. Calibration constants are derived  
during Calibration. These constants are used during runtime to maintain  
system accuracy. The Calibration constants are stored on the hard drive and  
reloaded at powerup. Complete system auto-calibration is achieved without  
any pots to adjust.  
The Selftest programs use the Selftest Assembly to test all circuits to their  
published specifications, and assure that all paths and modules are functional.  
The System Maintenance guide is written for troubleshooting to the module  
and circuit board level.  
The tester shall be disconnected from all power sources  
before it is opened for any adjustment, replacement,  
maintenance or repair. If service repairs performed under  
power become necessary, the repairs shall be performed  
by a skilled technician who is aware of the hazard  
involved.  
Series 2040 Test System  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance Overview  
Series 2040 Maintenance Manual V2.00  
Series 2040 Test System  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Selftest & Calibration  
Selftest & Calibration  
Series 2040 Test System  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Selftest & Calibration  
Series 2040 Maintenance Manual V2.00  
SELFTEST AND CALIBRATION OVERVIEW  
Calibration on the Series 2040 Functional Test System is completely  
automated. All calibration is based on high precision references built into the  
system Testhead and the Selftest Assembly. A high precision Temperature  
Compensated Crystal Oscillator (TCXO) on the Time Measurement System  
Board (TMS) is used as a frequency/time reference. A precision digital to  
analog converter (TDAC) contained within the Selftest Assembly is used as a  
voltage reference for system calibration.  
Purpose Of Calibration  
Calibration of the functional tester is necessary to obtain gain and offset terms  
for all signal paths that exist within the system, and to determine timing for the  
MDE’s (Measurement Display Electronics) sweeps, delays, triggers, and  
measurement mark. Since potentiometers are used in only a few instances  
where factory calibration is required, the majority of the calibration is done by  
taking measurements and calculating calibration terms which can be  
referenced at a later time. These terms are used by the hardware functional  
calls to account for gains, offsets, and timing discrepancies, and their use is  
transparent to the user.  
In all calibration programs, limits are set on what values any given calibration  
term may take on. If a calculated value is outside of the allowable range, a  
failure will result and be displayed/printed to the designated output device .  
When To Calibrate  
Selftest is the best guide to determine if calibration is required. The battery of  
Selftest programs is designed to test all functions of the 2040 to their  
specifications. If any failures occur during Selftest (especially borderline  
failures), calibration should be carried out before proceeding to troubleshoot  
the system.  
Running through Selftest periodically can tell you if calibration is or is not  
required. If you find that all Selftest programs run without failures, there is no  
need to recalibrate any portion of the tester. The recommended guideline is to  
calibrate every 3 months.  
Whenever a temperature change of more than five (5) degrees Celsius occurs  
or a Testhead board swap is made, the 2040 should be recalibrated.  
Series 2040 Test System  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Selftest & Calibration  
SELFTEST  
Selftest on the Series 2040 Test System is an automated procedure. The  
programs that comprise Selftest are organized under the Selftest Executive in a  
test menu called “Functional”. These programs exercise the hardware  
functions available to the user through the hardware functional calls. To  
perform Selftest, proceed as follows:  
1) Install the Selftest unit on the Patchboard assembly as shown  
below:  
A) Pull the locking handle towards the front of the unit until it  
is vertical.  
B) Slide the pins on the rear of the Selftest unit into the  
Patchboard and rotate the Selftest unit forward until it is  
vertical, and sets into the Patchboard.  
C) The ribbon assembly should be installed in the receptacle  
behind the Selftest, with the colored edge of the ribbon on  
the left side, while viewing the front of the unit.  
D) Rotate the locking handle back to it’s original position to  
lock the Selftest into place.  
2) Enter the Selftest Executive by selecting (double-clicking on) the  
“SelfExec” program from the Windows Explorer. The Selftest  
Executive menu will be displayed as shown on the next page.  
3) From the File menu, select Configure, and the small “Report  
Series 2040 Test System  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Selftest & Calibration  
Series 2040 Maintenance Manual V2.00  
Utilities” dialog box in shown on the  
next page will appear. This box contains  
three separate ways of reporting the  
data collected during the tests. Data  
may be stored to a file, displayed on the  
screen and/or printed.  
In addition, each of these three  
reporting functions can report All  
Data”, “Fail Data” or “No Data”. All  
Data” includes all of the pass or fail  
sequences from the test. “Fail Data”  
includes only information on the tests  
that failed. “No Data” disables that reporting function. To select, merely  
use the mouse to check or uncheck each reporting function. If a  
reporting device is unchecked, the options under that device appear  
“ghosted”, indicating that they are disabled. For diagnostic purposes,  
enable the “Printer” and “File” options, and select the “Fail Data”  
option. Under the “File” section, click the “Diagnostic File” check box.  
Disable the Window reporting functions.  
Series 2040 Test System  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Selftest & Calibration  
4) Once the type(s) of reporting is completed, click on “OK” to return  
to the Selftest Executive menu. Select “Functional” from the  
Selftest Executive screen, and the Functional menu will appear as  
shown below.  
The Functional test menu has been organized so that tests closely  
related are blocked together in groups. If only one test is to be run,  
use the mouse to select that particular test. If a related group of  
tests is to be run, they may be selected using only the mouse. For  
example, if the Measurement Display Electronics group is to be  
selected, place the mouse pointer on test #30 and hold the left  
button down, and drag the mouse down to test #38. The entire  
MDE group is now selected. If a random series of tests is desired,  
place the mouse pointer on the box under “Run sequence”. When  
the Windows text tool appears, click the left mouse button. At the  
blinking cursor, any sequence of tests can be entered. The test  
numbers, however, must follow this syntax:  
E’ (Execute) signifies the beginning of a sequence or acts as a  
separator between sequences.  
Series 2040 Test System  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Selftest & Calibration  
Series 2040 Maintenance Manual V2.00  
:” (Colon) signifies that the following number indicates the amount  
of iterations to perform of the previous sequence.  
,’ (Comma) shares the definition of ‘E’ as a separator between  
sequences.  
-’ (Dash) signifies inclusive or ‘through’.  
EXAMPLE: E1-3:5,6-100:2,1-5  
This sequence executes tests 1 through 3 five times, tests 6 through  
100 two times, and tests 1 through 5 once.  
If a single test or a group of selected tests is to be run once, 10  
times, or 100 times, use the mouse to select the appropriate button  
on the upper left. If the test or group is to be run at some other  
number, place the mouse pointer in the box to the direct left of the  
stop button, and the Windows text tool will appear in place of the  
mouse pointer. Click the left mouse button while texttool is  
displayed in the box, and a blinking cursor appears in the box. The  
user can now enter any number from 1 to 999 for the number of  
times a test should run.  
For diagnostic purposes, click  
the mouse button on the small  
box in the upper left corner of  
the menu next to All Tests”.  
Select the “1" button, and the  
program will run all of the  
functional tests once, while reporting the data to wherever it was  
directed from the ”Report Utilities" box. For diagnostics this path  
should include the “File” and “Printer” categories, and the “Fail  
Data” and “Diagnostic File” options.  
5) If any of the Functional programs fail, check these common  
problems:  
SELFTEST ASSEMBLY IMPROPERLY INSTALLED - Check to make  
sure the assembly is properly installed on the Patchboard. Also,  
Series 2040 Test System  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Selftest & Calibration  
check the Selftest assembly’s cable for damage.  
UUT POWER SUPPLIES OFF - Make sure the UUT Power Supplies  
are on.  
TESTHEAD POWER SUPPLY FAULT - Check the Testhead Power  
Supply Controller’s indicator LEDs. If any of the LEDs are not lit,  
reset the system and recheck.  
MISALIGNED PINS ON PATCHBOARD RECEIVER - Remove the  
Selftest Assembly and carefully inspect the alignment of the  
Patchboard receiver pins. Adjust any pins that are misaligned,  
remount the Selftest Assembly, and re-test.  
After the common problems have been eliminated, a failure in  
Functional Selftest can indicate a hardware problem. Since a failure  
in Functional Selftest could also be caused by calibration,  
recalibrate the unit and run Functional Selftest again. If calibration  
or Functional Selftest fails now, a hardware problem most likely  
exists. Refer to the Selftest Failure Analysis on the next two pages.  
Series 2040 Test System  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Selftest & Calibration  
Series 2040 Maintenance Manual V2.00  
Selftest Failure Analysis  
Test Name  
#1  
#2  
#3  
1)  
2)  
3)  
4)  
SELF_dig-f  
th_config_tst  
ams_dig_f  
ams_mode_f  
ams_sig3_f  
Selftest  
TH Cont/CPU TH P/S Cont  
Testhead configuration does not match the stored configuration.  
AMS  
AMS  
AMS  
TH Cont/CPU TH P/S Cont  
MSP/ISO  
MSP/ISO  
ASB  
TH P/S Cont  
5)  
6)  
7)  
8)  
9)  
arly_dig_f  
arly_f  
afet_rft_f  
Arly/AFET  
Arly/AFET  
AFET  
TH Cont/CPU TH P/S Cont  
RMUX  
RMUX  
AMS  
AMS  
10)  
11)  
12)  
13)  
14)  
15)  
16)  
17)  
18)  
19)  
20)  
21)  
22)  
23)  
24)  
25)  
26)  
27)  
28)  
29)  
30)  
31)  
32)  
33)  
34)  
35)  
36)  
37)  
38)  
39)  
40)  
41)  
42)  
43)  
44)  
45)  
mrly_dig_f  
mrly_f  
MRLY  
MRLY  
Selftest  
RMUX  
TH Cont/CPU  
AMS  
rmux_dig_f  
rmux_f  
rmux_prt_f  
RMUX  
RMUX  
RMUX  
TH Cont/CPU TH P/S Cont  
AMS  
AMS  
Selftest  
MDE  
arb_mem_f  
arb_brst_f  
arb_ext_f  
arb_freq_f  
arb_mon_f  
arb_ref_f  
da_ref_f  
ASB  
ASB  
ASB  
ASB  
ASB  
ASB  
ASB  
TH Cont/CPU TH P/S Cont  
AMS  
AMS  
AMS  
AMS  
AMS  
AMS  
MDE  
MDE  
MDE  
MSP/ISO  
MDE  
MSP/ISO  
mde_dlmd_f  
mde_dltm_f  
mde_expm_f  
mde_freq_f  
mde_mrkp_f  
mde_msmk_f  
mde_pmes_f  
mde_tgfl_f  
MDE  
MDE  
MDE  
MDE  
MDE  
MDE  
MDE  
AMS  
AMS  
TMS  
TMS  
TMS  
TMS  
TMS  
TMS  
TMS  
MDE  
MDE  
AMS  
AMS  
AMS  
AMS  
AMS  
AMS  
AMS  
TMS  
TMS  
mde_tglv_f  
tms_dig_f  
tms_evnt_f  
tms_test_f  
TMS  
TMS  
TMS  
TH Cont/CPU TH P/S Cont  
MDE  
MDE  
AMS  
TH P/S Cont  
dio_f  
DIO  
Selftest  
TH P/S Cont  
Series 2040 Test System  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Selftest & Calibration  
46)  
47)  
48)  
49)  
50)  
51)  
52)  
53)  
54)  
55)  
56)  
57)  
58)  
59)  
60)  
61)  
62)  
63)  
64)  
65)  
66)  
67)  
68)  
69)  
70  
dio_ext_f  
DIO  
Selftest  
TH P/S Cont  
iamp_dig_f  
inst_f  
amp_cmrr_f  
isoamp_f  
INST  
INST  
INST  
MSP/ISO  
TH Cont/CPU TH P/S Cont  
ASB  
RMUX  
ASB  
AMS  
AMS  
AMS  
pwr_mon_f  
ptest_f  
th_iso_f  
Power Supply  
UUT Controller AMS/RMUX  
Scope Connected?  
TH P/S Cont  
MSP/ISO  
Selftest  
Internal coax short?  
msp_ser_f  
MSP  
TH Cont/CPU TH P/S Cont  
ocio_dig_f  
ocio_f  
ocio_rail_f  
OCIO  
OCIO  
OCIO  
TH Cont/CPU TH P/S Cont  
Selftest  
Selftest  
TH Cont/CPU  
TH Cont/CPU  
adio_f  
adio_ext_f  
ams_int_f  
ADIO  
ADIO  
AMS  
Selftest  
Selftest  
MDE  
TH Cont/CPU  
TH Cont/CPU  
TH Cont/CPU  
dms_dig_f  
dms_mem_f  
DMS  
DMS  
Selftest  
TH Cont/CPU  
TH Cont/CPU TH P/S Cont  
71)  
72)  
psc_dig_f  
PSC  
Any other Trigger Matrix board.  
Series 2040 Test System  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Selftest & Calibration  
Series 2040 Maintenance Manual V2.00  
CALIBRATION  
Whenever a Testhead board is  
replaced, or any time that Selftest  
determines a failure, a calibration  
must be performed. If a unit  
passes the complete Functional  
Selftest routine, a calibration is  
not necessary. When Calibrate is  
selected from the Selftest  
Executive, the screen shown to  
the right is displayed.  
The Calibrate menu operates  
under the same basic rules as the  
Functional menu. For installation  
purposes, use the mouse to select the All Tests” box, and select the “1"  
button for the number of times to be run. (See page #6). Ensure that the  
Printer and Fail Data are selected, so that any failures will be printed  
automatically.  
The following is a list of the calibration programs located in the Calibrate  
menu in the Selftest Executive, with the three most likely boards or assemblies  
that could cause failures:  
Test Name  
#1 AMS_RMUX_c AMS  
#2 AMS_SIG3_c  
#3 AMS_DIFF_c  
#4 DMS_c  
#5 ARB_v_c  
#6 DA_v_c  
#1  
#2  
#3  
RMUX  
MSP/ISO  
RMUX  
MSP/ISO  
RMUX  
ICAM  
AMS  
AMS  
DMS  
ASB  
MSP/ISO  
MSP/ISO  
AMS  
AMS  
ASB  
#8 DIO_c  
#9 ADIO_c  
DIO  
ADIO  
TH CONT/CPU Selftest  
TH CONT/CPU Selftest  
#10 ADIO_dac_c ADIO  
#12 MDE_c MDE  
#13 MDE_TGDC_c MDE  
AMS  
TMS  
TMS  
RMUX  
AMS  
AMS  
#15 POWER_c  
#17 ISOAmp_c  
UUT Cont  
MSP/ISO  
AMS/RMUX Selftest  
AMS RMUX  
Series 2040 Test System  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Selftest & Calibration  
Any type of failure during calibration could indicate a hardware problem. If a  
failure during Functional was the reason that a calibration is being run, refer to  
the Selftest Failure Analysis sheet on page 8 of this section.  
Series 2040 Test System  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Selftest & Calibration  
Series 2040 Maintenance Manual V2.00  
Series 2040 Test System  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
Diagnostics  
Series 2040 Test System  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
DIAGNOSTIC TOOLS  
SelfTest EXaminer (STEX) for Microsoft Windows 95  
Overview  
The STEX program may be used to help graphically interpret the  
results of any diagnostic output file generated by Digalog System’s  
SelfExec program run on the Series 2040 testers. STEX uses the results of the  
diagnostic output file to show the configuration of the tester and evaluate  
possible problems with the tester resources based on failed test data and/or  
system errors during testing. STEX will use these results to ‘decide’ what the  
top three most likely resource problems are in the tester, based on Selftest  
output information. These results will be shown graphically on the STEX  
interface, highlighting and identifying the potentially problematic resources in  
the tester.  
STEX can be very useful in aiding an operator who is troubleshooting possible  
problems in the Digalog testers by providing a graphical depiction of the  
location of potentially faulty tester resources. These resources include all  
Testhead boards, the Testhead power supplies, the variable power supplies  
and their controller modules, the computer with its Testhead controller card,  
and the Selftest unit itself.  
Creating the Diagnostic File for STEX  
To use STEX to analyze Selftest data, the proper information must be recorded  
in the Selftest output file first. A new user selection under SelfExec’s output  
configuration has been added to provide this information. To obtain a  
diagnostic file for use in STEX, follow these steps  
when running SelfExec:  
1) Under the ‘File’ menu, select  
Configuration’.  
2) On the configuration form under ‘File  
Options’, select either ‘Fail Data Only’ or All  
Data’ and select ‘Diagnostic’.  
3) Run the desired calibration and function  
tests.  
Series 2040 Test System  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
Selection of the ‘Diagnostic’ option for file output will create a file with all the  
information required by STEX to analyze the system. This information  
includes the tester type being run, a listing of the Testhead resources as  
recorded in the ‘resource.ini’ file or the Registry, and a listing of all tests run  
along with any problems encountered during the tests. These results will be  
written into a file in the ‘digalog\bin\’ directory. The filename is the date on  
which the test was run with the file extension ‘.dxx’, where ‘xx’ is an  
incrementing number starting at ‘01’ and going through ‘99’. For example, if  
the diagnostic file was written on January 20, 1997, the file name would be  
‘jan2097.d01’. If a second file were created on the same day, its name would  
be ‘jan2097.d02’, and so on. The first SelfExec file written each day would  
carry the extension ‘.d01’. If SelfExec’s file option is not set to ‘Diagnostic’, the  
file extension would be ‘.001’, ‘.002’, and so on. These files do not contain  
the configuration information required by STEX and will not be accepted as  
proper files for analysis by STEX.  
Analyzing Selftest Data  
When starting the STEX program, the graphics show the Testhead as being  
empty. Under the menu item ‘File’ are the selections: Open for EXam, View  
Log File, Report Details, Close, and Exit.  
Open for EXam - Selecting this menu item opens up a  
file selection window showing the files available in the  
‘digalog\bin\’ directory with the ‘.dxx’ file extension.  
Highlight the desired filename and Click on the ‘Open’  
button to select for analysis under STEX. STEX will then  
open that file, and examine it to see if it is a valid file for  
this program. If the file is not valid for STEX, a message  
box will inform the user of this, and the graphics will  
return to their default condition. If it is valid, the tester  
resource information will be read in for analysis. The boards listed in the  
Registry will be placed in the proper slots in the Testhead graphic and the file  
will be inspected for any fail data from the tests listed. If fail data or system  
errors are found, the program will analyze those results using information  
about the tests located in the STEX database — ‘digalog\include\stex.mdb’.  
The program will then use this information to calculate the three most likely  
tester resources that could cause the problems found in this diagnostic file.  
Series 2040 Test System  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
All tester resources, by default, are either gray-colored or on gray background.  
The one to three resources calculated to be most likely causing the fault(s) will  
be highlighted in red. Also, the ‘Results:’ box on the lower portion of the form  
displays the information regarding these three tester resources. If the resource  
is a board in the Testhead, an arrow with a number 1-3 according to its  
relative importance (1 being greatest) will point to the location of the board.  
An information box with the same number will display the board’s  
description, Digalog part number, and serial number (if this information is  
available in the diagnostic file.) If the tester resource is not a board in the  
Testhead, the resource’s graphic will be highlighted in red with a label  
indicating which ‘Choice #’ it is in relative importance. The information box  
with the ‘Choice #’ information will have the resource’s description and other  
optional information regarding that resource.  
If the SelfExec test ‘TH_config_test’ has logged fails in the current diagnostic  
file, a message box will be shown informing the user that this fault exists.  
Also, the ‘Results:’ box will display the message: “Config problems found” on  
a red background. The configuration of the Testhead will be displayed, but no  
specific tester resources will be tagged as being at fault as a result of this file’s  
analysis. The test ‘TH_config_test’ examines and compares the contents of  
the Testhead to the recorded listing in the Registry. If a problem is found with  
this test, it cannot be automatically known where the source of the problem is  
without further investigation. Therefore, the user should refer to the Tester  
Resource Manager (TRMAN) section of this manual to proceed in  
troubleshooting this problem.  
Series 2040 Test System  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
If any failure(s) or problem(s) are found in a valid diagnostic file, the menu item  
‘Report Details’ will be enabled to show some details about those tests. See  
item ‘Report Details’ for more information.  
If the diagnostic file is examined, and no fails or other problems were found  
with this test, the ‘Results:’ box will display the message: “No Problem  
Found” on a green background. The proper configuration for the Testhead  
will be displayed, but the menu item ‘Report Details’ will be disabled, since  
there are no failing tests on which to display information.  
View Log File - Selecting this menu item opens a file selection window  
showing the available files, similar to the ‘Open for EXam’ item. Selecting a  
filename here will start the Microsoft Windows text editor program ‘WordPad’  
and open the diagnostic file for viewing. Here, individual test results may be  
viewed to provide further possible insight into any problems with the tester.  
Viewing the individual failures of a test may aid in troubleshooting an  
individual board. Examining any system error messages and referring to  
documentation for the source and/or meaning of the error may also help in  
resolving the problem with the tester resources. While viewing this file, the  
user may also elect to print the file contents.  
IMPORTANT!! The file being examined SHOULD NOT be edited. Changing  
any information in these files could render the information inaccurate or  
useless for diagnosing any problems with the tester.  
Report Details - This menu item is only available while there is a valid  
diagnostic file with fail information opened for examination. ‘Report Details’  
is a listing of the individual Selftest programs that show at least one fail or  
system error during its execution.  
Selecting this menu item will  
bring up a separate window  
with a list of the failed test  
programs. This list gives the  
name of the Selftest which had  
the problem, the slot number of  
the board being examined by  
that test, and whether the problem with the test was a test fail or a system  
Series 2040 Test System  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
error. If a test listed did not have any slot information listed in the results, the  
slot will be listed as ‘Unknown’. To view the details of each individual fail or  
system error number from the test results, use the ‘View Log File’ menu item.  
Refer to the Selftest documentation to aid in interpreting these test results.  
Close - Selecting this menu item closes the currently selected diagnostic file  
and clears all result information from the program’s windows. This clears the  
on-screen Testhead board configuration and removes any individual test  
results from the ‘Report Details’ window. This menu item is provided only for  
the convenience of the user. It is not mandatory to ‘Close’ one test before  
opening another for exam. However, it may be used to provide an obvious  
separation in the examination of different test result sets.  
Exit - Selecting this menu item closes the current diagnostic file, clears  
configuration information from the program, and closes and exits the program.  
Series 2040 Test System  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
TRMAN (Tester Resource Manager)  
The Tester Resource Manager is used to track and manage the tester resources  
including all the boards in the Testhead and the UUT power supplies.  
Information about these resources can be automatically generated or  
manually defined, and is used to generate a Patchboard Interface Map and  
define the pin locations of these resources at the Patchboard. This information  
can be saved to a project specific file called resource.ini and can be used by  
other Digalog System’s applications.  
Specifically, the software is capable of printing out a “Patchboard Map”  
containing Patchboard pin mnemonics by either automatically interrogating  
the tester for its resources, or by asking the programmer to define the tester’s  
resources. When the configuration is performed manually, it allows the  
programmer to configure additional resources beyond what the tester  
physically contains. In this manner, a programmer has the additional resources  
to generate programs and fixtures for any tester.  
Menu Bar  
File Menu - The TRMAN File menu is similar to the standard Windows file  
Series 2040 Test System  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
menu. It has selections for a New  
configuration, Load an existing  
configuration, Save a configuration, Save  
an existing configuration As another file,  
and Exit. The Print Utility as shown  
below will be discussed later in this  
section. For the purposes of this  
explanation, the file being used for  
illustrations is the standard resource.ini  
file in the \Digalog\include\ directory.  
Options Menu - When the Options menu is  
selected, a pull-down menu will be displayed  
prompting for a choice of the following: Auto-  
matic generation, Link to tester, Manual Gen-  
eration, View pin Definitions, and View  
Testhead configuration. Each of these options,  
along with the Print utility from the File menu,  
will be briefly discussed.  
Link to Tester - This item will toggle the link between the application  
and a tester. When checked, the application assumes a Testhead is  
present and enables Automatic generation. When using Tester  
Resource Manager on a development computer that is not connected  
to a Digalog tester, this item should never be checked as it could lead  
to corruption of certain memory locations.  
Automatic Generation - If Automatic Generation is selected, the  
program scans the tester for its resources, and then updates the map of  
the Patchboard as shown on the next page. Tester resources required  
to generate the Patchboard map plus any other tester resources can  
also be saved to a resource file on the hard drive (resource.ini) stored in  
the Registry, or printed out.  
All programmers writing code for the system should be familiar with  
this map, since it is the actual physical configuration of the  
Patchboard.  
Series 2040 Test System  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
Manual Generation - If Manual generation is selected, the screen  
below will be displayed. Note that the Power Supply Distribution  
board is always in slot 0 and therefore does not show in the Testhead  
Series 2040 Test System  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
configuration table. From this utility, any system configuration may be  
generated. It is the users responsibility to make sure the generated  
configuration is valid.  
The Programmable variable power supplies (maximum of five) may be  
defined in terms of maximum voltage and current once a  
Programmable Power Supply is selected (up/down button). With Volts  
or Amps selected, the jog shuttle located to the right modifies the  
selected value. The option buttons under the jog shuttle control  
determine what position in the serial loop that the supply being  
defined occupies. However, the variable supplies must be filled in a  
contiguous manner from #0 to #4. The spin control on the upper left  
corner of the dialog is used to determine which programming channel  
is being configured.  
If a GPIB or HPIB power supply is being defined or added, the channel  
and type can be selected and the upper right of the dialog changes to  
allow the programmer to select a power supply type, a GPIB/HPIB  
Device number, and what Relay Control board will be used. The spin  
control on the middle left of the dialog (Patchboard Connection) is  
used to define which of the five Patchboard connections are connected  
to which supply, or if a GPIB/HPIB supply will use an external output  
(i.e. the supply’s output does not physically appear on the Patchboard.)  
The rest of the Testhead is displayed in table format by slot number.  
The Description and Board Number drop-down menus are directly  
linked to, and will modify the table. The Clear All button will clear all  
configuration items.  
Board Codes are specific identifiers for the particular type of board  
selected for that slot. Board Numbers are used to define the resources  
of that particular board. For example, if there are two of the same type  
of board in the system, board numbers zero and one, the resources of  
board one will be numbered over and above those of board number  
zero. In other words, for a Relay Multiplexer Board containing sixty-  
four channels, board zero channels would be labeled zero through  
sixty-three and board one channels would be labeled sixty-four  
through one hundred and twenty-seven.  
Series 2040 Test System  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
The table is directly linked to the slot drop-down menu, therefore any  
slot changes in either place will be reflected in the other.  
View Testhead Configuration - This utility displays the entire Testhead  
configuration in an organized format. An example of a typical Testhead  
configuration is shown below.  
Series 2040 Test System  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
View Pin Definitions - This utility from the  
Options menu explains the mnemonics  
used in the Patchboard map as shown to  
the right.  
View Power Supply Configuration - If this  
option is selected, a grid is displayed  
showing all of the power supplies present,  
along with the type, name, device  
designation, and Patchboard connection  
for each supply as shown below.  
Print Utility - When this utility is selected from the File menu, a small  
inputbox is displayed prompting for a specific serial number for the  
printout to be generated. If a specific tester’s configuration is to be  
generated, enter the serial number in the textbox. If not, merely select  
the OK command button, press <Enter> on the keyboard, or select  
Cancel. After this inputbox is satisfied, select an individual printout to  
be generated, or select all three options.  
Series 2040 Test System  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
TESTHEAD - PATCHBOARD - UUT P/S BUILT-IN DIAGNOSTICS  
When a malfunction is suspected in either the Testhead Internal or  
Patchboard Power Supply system, the LED status indicators should be  
examined. The top LEDs are labeled for their respective power supplies as  
shown on page 14. The legend for the four lower LEDs is shown below. In  
case of a power supply malfunction, the power supply that has shut down first  
will be indicated by its respective LED being extinguished.  
If you are diagnosing the case of no power at the Patchboard, and the status  
indicators, the lower four LEDs, are all lit, then an over/under voltage fault has  
occurred in a Patchboard power supply. The power supply causing the fault  
will have it’s respective LED extinguished. Power could also have been  
removed from the Patchboard without generating a fault condition. In this  
case, the reason that the power has been removed is indicated by one or two  
of the lower four status LEDs being extinguished.  
CASE  
PROX  
PBON  
Ext.Fault  
Reset  
Power Call  
TCLEAR  
Patchboard removed  
on  
on  
off  
off  
off  
on  
on  
on  
on  
on  
off  
on  
UUT Power Fault  
on  
on  
off  
on  
When a power supply cannot be restored by resetting the unit, the circuit  
breaker for the offending supply should be checked for continuity. A graphic of  
the circuit breakers on the edge of the Testhead Power Supply Controller  
board is shown on page 14 of this section. If any of these breakers continually  
trips, the associated power supply must be replaced.  
Series 2040 Test System  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
Breakers and Test Points  
LED Indicators  
TP17  
+48VDC TH  
DC1 -  
+ 5VDC TH  
+ 5VDC TH  
+ 48V TH  
TPA17  
DC1 +  
TP16  
+ 48V TH  
TPA16  
GPIB 3  
GPIB 2  
GPIB 1  
GPIB 0  
TP14  
DC1+  
DC1-  
DC1+  
DC1-  
TPA14  
TP13  
TPA13  
- 40VDC (ICAM)  
+ 40VDC (ICAM)  
- 5.2VDC TH  
TP19  
- 40VDC TH  
TPA19  
TP18  
+/- 40VDC TH  
+ 5VDC TH  
- 19VDC TH  
+ 19VDC TH  
- 15VDC TH  
+ 15VDC TH  
+ 40VDC TH  
TPA18  
TP8  
- 19VDC TH  
- 19VDC TH  
+ 19VDC TH  
- 15VDC TH  
+ 15VDC TH  
- 5.2VDC PB  
TPA8  
TP7  
+ 19VDC TH  
+ 5VDC TH  
TPA7  
TP6  
- 15VDC PB  
+ 15VDC PB  
+ 5VDC PBF  
HP Fault  
TPA6  
TP5  
+/- 15VDC TH  
Patchboard Power  
TPA5  
TP4  
TPA4  
PROX  
PBON  
Ext. Fault  
Reset  
TP3  
Common  
- 15VDC PB  
+ 15VDC PB  
TPA3  
TP2  
K2  
+ 5VDC  
PB  
K1  
+/- 15VDC  
PB  
TPA2  
TP1  
+ 5VDC PB  
TPA1  
Series 2040 Test System  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
FAULT LOOP  
The Series 2040 Test System incorporates a “fault loop” for troubleshooting  
and self-diagnostics. The cable between the UUT Power Supply Controller  
and the Testhead, besides providing a path for the power supply output,  
contains the fault loop. If the fault loop is broken, it will cause the controller  
to shut down and report an error to the computer (the next time the computer  
tries to give it a command). The fault loop is broken by the controller itself  
whenever the controller detects an error. This signals to the other controllers  
that they too should shut down. If a controller for a power supply channel is  
unresponsive and the “Ext.Fault” LED on the Testhead Power Supply  
Controller is on, a communication problem exists between the CPU and the  
controller. If the LED is out, the fault loop is probably broken.  
The power supplies may be isolated for diagnostics as shown in the following  
pages. If a supply is isolated and fails to power up, it is most likely the  
problem. If a supply is isolated and powers up, it can b eliminated as the  
cause of the failure.  
Series 2040 Test System  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
Series 2040 Test System  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
Series 2040 Test System  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
Series 2040 Test System  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
GPIB/HPIB POWER SUPPLIES  
The Series 2040 Test System may also contain one or more GPIB/HPIB Power  
Supplies. These supplies communicate with each other and the computer via  
an IEEE-488 bus. In addition, they use a Fault/Inhibit loop for voltage/current  
protection and this loop also connects to the Testhead Power Supply  
Controller.  
Some of these supplies may have more than one output channel. In these  
situations, each separate output of each GPIB/HPIB supply is assigned its own  
channel or unit number. Programming is accomplished using Digalog’s  
PowerUUT functional call. Over/under voltage/current limits for these supplies  
and channels are listed in the “resource.cat” file. For programming purposes,  
these power supplies act in the same manner as the Digalog UUT Product  
Power Supplies.  
Troubleshooting Hewlett-Packard Power Supplies  
The fault/inhibit loop for the Hewlett Packard power supplies links each HP  
power supply serially with each other and with the PSC board. It is completely  
independent of the UUT P/S controllers’ (external) fault loop and the fault  
loop for each non-HP GPIB power supply.  
The PSC board will recognize a fault generated by the HP fault/inhibit loop  
only if the registered power supply configuration (set up by TRMAN) contains  
at least one HP power supply whose “resource.cat” entry has its “Fault  
Capable” property set to “1.”  
The “HPF” LED on the PSC board will turn off whenever an HP power supply  
has either detected an internal error condition, such as overvoltage, or has  
been “faulted” by the PSC board, but only after a HP power supply has been  
programmed following a Testhead power-up or the TClear event.  
The fault signal driving the HPF LED on the PSC is not latched. Also, if the  
reset jumper shunt on the PSC board is installed, only the Proximity-Switch  
LED will turn off when the Patchboard fixture handle is disengaged. It will  
then turn back on when the handle has re-engaged. The PSC board will drive  
its HP fault/inhibit loop connection low every time that it detects a fault from  
any source whether or not any HP power supplies are registered or even  
present.  
Series 2040 Test System  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
For a system containing “n” number of HP power supplies, the PSC board  
drives the +Inhibit (input) line of HP P/S #n low relative to its -Inhibit (input)  
line. When this power supply detects that its +Inhibit input signal is low, then  
it drives its +Fault (output) signal low relative to its -Fault (output) signal.  
These fault output lines are connected to the inhibit input connections of HP  
P/S #(n-1). Upon its detection that HP P/S #n has driven its +Inhibit (input)  
signal low, HP P/S #(n-1) will then “fault” HP P/S #(n-2) using its fault (output)  
connections. Thus, each HP P/S in the fault loop will “fault” the power supply  
preceding it in the loop all the way to HP P/S #0. The Fault (output) lines of  
HP P/S #0 connect to the PSC board, which allows that power supply to  
signal the PSC board that an error has been detected by one of the HP power  
supplies.  
Upon detection of an internal error condition, such as overvoltage, or the  
external inhibit (input) signal, the HP power supply will program its output(s)  
to 0.0V @ 0.0A, disable its output(s), and assert its fault (output) signal.  
If an HP power supply detects an internal error condition, it will then drive its  
+Fault (output) line low to signal the next HP power supply positioned below  
it in the loop. Eventually, the “fault” signal will reach HP P/S #0, and it, in  
turn, will trigger the PSC board. The PSC board will then latch HP P/S fault  
condition and drive all of its fault output connections, including that for the  
HP power supplies. This will “fault” all of the HP power supplies positioned  
after the one which detected the internal error condition. The initial faulting  
HP power supply will also have its +Inhibit (input) connection driven low by  
the power supply following it.  
To determine which HP power supply detected an internal error condition,  
press the “Local” button and then the “Prot” button on each HP power  
supply. All of the HP power supplies that were “faulted” by their following  
brethen via their inhibit (input) lines will display the letters “RI,” which stands  
for “Remote Inhibit,” in its LCD display. The single HP power supply that  
displays something other than “RI” in its LCD display is the power supply that  
detected an internal error condition. Consult its operating manual to  
determine the cause of the internal error condition.  
If some or none of the HP power supplies are disabled via their fault/inhibit  
loop connections when the PSC board detects a fault from some other source,  
such as the proximity switch, proceed with the following troubleshooting  
procedure.  
Series 2040 Test System  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
1) Verify on the relay controller board (0000-5X44) that pin #3 relative  
to pin #4 of connector P3 (PSC Port) is low or at a logic “0.” If not, a  
problem exists on the PSC board, in the fault/inhibit cable from the  
PSC board to connector P3, or with connector P3 .  
2) Otherwise, verify that pin #3 relative to pin #4 of connector P5  
which mates to the first relay disconnect board is low or at a logic “0.”  
If not, a problem exists on the relay controller board.  
3) Otherwise, on the first relay disconnect board, verify that pin #3  
relative to pin #4 of connector P1 (which mates to the relay controller  
board) is low or at a logic “0.” If not, a problem exists either with  
connector P1 or else with connector P5 on the relay controller board.  
4) Otherwise, verify that pin #3 relative to pin #4 of connector P3  
(which mates to the next relay disconnect board) is low or at a logic  
“0.” If not, a problem exists between connectors P1 and P3 on the  
relay disconnect board.  
5) Otherwise, on the second relay disconnect board, verify that pin #3  
relative to pin #4 of connector P1 (which mates to the first relay  
disconnect board) is low or at a logic “0.” If not, a problem exists either  
with connector P1 or else with connector P3 on the first relay  
disconnect board.  
6) Repeat steps #4-5 for the second and all subsequent relay  
disconnect boards up to and including the last relay disconnect board.  
7) Jumper shunts #7-8 (which connect the corresponding inhibit and  
fault signals) on the last relay disconnect board must be installed. Verify  
continuity from pin #3 of connector P1 through jumper shunt #8 to  
pin #4 of connector P2 (which connects to the fault/inhibit cable for  
the HP power supply). Verify continuity from pin #4 of connector P1  
through jumper shunt #7 to pin #3 of connector P2.  
8) Otherwise, verify that pin #4 relative to pin #3 of connector P2 is  
low or at a logic “0.” If not, a problem exists on the relay disconnect  
board, in its fault/inhibit cable, or with the HP power supply’s inhibit  
input pins. Verify the continuity of each wire in the fault/inhibit cable,  
Series 2040 Test System  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
and verify that each wire is connected to the proper terminal on the HP  
power supply. The HP power supply may be configured to use its  
inhibit input pins as simply digital I/O pins. Consult its operating  
manual to determine the correct position of the internal configuration  
jumper shunt for “Fault/Inhibit” operations, and then remove its cover  
to confirm the shunt’s proper placement.  
9) Otherwise, verify that an arrow exists above the word “Prot” on the  
LCD display. If not, the HP power supply itself is faulty.  
10) Otherwise, verify that pin #1 relative to pin #2 of connector P2 on  
the relay disconnect board is low or at a logic “0.” If not, a problem  
exists in the HP power supply, in the fault/inhibit cable for this supply,  
on the relay disconnect board for this supply, on the relay disconnect  
board for the previous supply, in the fault/inhibit cable for the previous  
supply, or in the previous HP power supply itself.  
11) Otherwise, verify that pin #1 relative to pin #2 of connector P1  
(which mates to the previous relay disconnect board) is low or at a logic  
“0.” If not, a (continuity?) problem exists on the relay disconnect board.  
12) Otherwise, on the previous relay disconnect board, verify that pin  
#1 relative to pin #2 of connector P3 (which mates to the next relay  
disconnect board) is low or at a logic “0.” If not, a problem exists either  
with connector P3 or else with connector P1 on the next relay  
disconnect board.  
13) Repeat steps #8-12 for each previous relay disconnect board up to  
and including the first relay disconnect board.  
14) Otherwise, on the relay controller board, verify that pin #1 relative  
to pin #2 of connector P5 (which mates to the first relay disconnect  
board) is low or at a logic “0.” If not, a problem exists either with  
connector P5 or else with connector P1 on the first relay disconnect  
board.  
15) Otherwise, verify that pin #1 relative to pin #2 of connector P1  
(PSC Port) is low or at a logic “0.” If not, a problem exists on the relay  
controller board, in the fault/inhibit cable from the PSC board, or on  
Series 2040 Test System  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Diagnostics  
the PSC board itself.  
Note: If there are no wires in the fault/inhibit cable of the HP power  
supply for the fault and inhibit signals (pins 1-4 on connector P2), then  
jumper shunts #9-10 (Fault/Inhibit Bypass) must be installed.  
Series 2040 Test System  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Diagnostics  
Series 2040 Maintenance Manual V2.00  
Series 2040 Test System  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Certification  
Certification  
Series 2040 Test System  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Certification  
Series 2040 Maintenance Manual V2.00  
X
Y
Z
10MHz  
10 M H Z E XTC LEKXTE VE N T  
TDAC  
GPIB  
G PIB  
P R IN TER  
ETH E R N ET  
C OM 1  
K EY BO AR D  
M O D EM  
Series 2040 Test System  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Certification  
CERTIFICATION  
System certification ensures that the system standards that are used for Selftest  
and calibration are accurate, and can be traced to the National Institute of  
Standards and Technology (NIST). There are two elements of the Series 2040  
Functional Test System that need to be certified:  
1) The TDAC high precision digital to analog converter on each  
Selftest unit.  
2) The TCXO precision oscillator on the Time Measurement System  
(TMS) in the 2040 Testhead.  
To perform a certification, the following equipment is necessary:  
1) A Digalog Calibration Station  
2) A Digalog Selftest assembly (Mounted on the 2040)  
3) A printer (if a hard copy certificate is desired)  
The Calibration Station contains a precision digital multimeter and a precision  
counter/timer. Both instruments are equipped with a GPIB interface so that  
the certification process (in terms of instrument setup and measurement  
processing) is completely automatic.  
Setup  
A number of connections between the 2040 and the Calibration Station must  
be made before continuing. These are:  
1) Installation of the Selftest assembly. (See page 2-1.)  
2) Power connections for the Calibration Station.  
3) GPIB IEEE-488 communication link between the 2040 and the  
Calibration Station.  
4) Connections between the TDAC port on the Selftest assembly and  
the multimeter.  
5) Connections between the 10MHz output of the TMS in the  
Testhead and the counter/timer.  
Using the interconnect on the left page, hook a BNC cable from the 10MHz  
connector on the back panel of the 2040 Console to the A” input of the  
Frequency counter of the Calibration Station. Using another BNC cable,  
connect the TDAC signal from the Selftest unit to the Multimeter on the  
Series 2040 Test System  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Certification  
Series 2040 Maintenance Manual V2.00  
Calibration Station. Finally, connect the GPIB port from the 2040 console to  
the GPIB port on the Calibration Station.  
To run the program select the  
certification icon from the Atart/  
Programs/Digalog program group,  
and the screen to the right will  
appear prompting for the type of  
calibration. Select the TDAC, and  
the screens in the right column on  
both this page and the following  
page will appear in the order  
shown.  
The first screen prompts for the  
user’s name, the serial number of  
the 8840 Fluke multimeter in the  
Calibration Station. This number can be found on a Digalog tag inside the  
Calibration Station next to the  
meter. It also prompts for the  
Selftest Assembly serial number,  
and the serial number of the  
tester itself.  
When the information is  
entered, the last setup screen  
will appear requesting a  
verification of the information  
that was entered. If any of the  
information is incorrect, it can  
be modified or changed at this  
time.  
The operator’s name, Selftest  
serial number, and Fluke 8840A  
meter serial number must be  
entered and verified before the  
unit can be certified. Click on the “Certify” button and the program checks  
the TDAC signal. Test results and a pass/fail condition are displayed in the  
Series 2040 Test System  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Certification  
screen shown below.  
When this portion of the certification is complete, the TCXO frequency must  
be checked. The procedure for the frequency certification is identical, except  
for the final screen displaying the frequency counter. The user should wait  
until the frequency is stable, and then click on certify to start the test. Again,  
the results and a pass/fail condition are displayed.  
For the portion dealing with the TDAC, the measured value must not vary  
from the programmed value by more than the designated limits or a failure  
will result. If this happens, a manual adjustment must be made. Contact  
Digalog Systems should this occur.  
Series 2040 Test System  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Certification  
Series 2040 Maintenance Manual V2.00  
If the TCXO fails the preset calibration limits, it will also require a manual  
adjustment. Again, contact Digalog Systems should this occur.  
The procedure for the Selftest PPS current feedback circuit is almost identical  
to the first two certification procedures. Follow the instructions displayed on  
the screen.  
The fourth procedure certifies the ICAM Selftest current measurement  
circuitry, and the last certification screen is shown below.  
Connect the ICAM Certification cable from the white connector on the ICAM  
Selftest board to the Fluke 8840 Multimeter.  
The ICAM IMON measurement certification tests the circuitry over six  
different ranges with the results displayed to the right of the dialog. The result  
of the current certification is also displayed. If any of these readings are out of  
range, the small box adjacent to the result will appear red instead of green.  
Series 2040 Test System  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Module Repair and Replacement  
Module Repair & Replacement  
Series 2040 Test System  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Module Repair and Replacement  
Series 2040 Maintenance Manual V2.00  
REPAIR AND REPLACEMENT PROCEDURES  
Upon determining that a part is defective, the customer should:  
1) Contact Digalog Systems at (414) 797-8000 to obtain an RMA (Return  
Materials Authorization) number and be prepared to supply the part number  
and serial number of the defective part. If the part is not covered under  
warranty or a support contract, the customer must provide a method of  
payment for the repairs.  
2) Return the defective part and reference the RMA number. All equipment  
being returned to Digalog Systems should be shipped prepaid and insured at  
the customer’s expense.  
3) All circuit boards to be returned to Digalog Systems must be shipped in  
static shielding bags and packed in anti-static packaging material. Digalog  
Systems will pay the return freight, within the continental United States on a  
warranty issue. Digalog Systems is not liable for consequential shipping  
damages.  
4) Out-of-warranty repairs will be done on a time-and-materials basis (or may  
be covered under a Support Agreement.)  
Series 2040 Test System  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Module Repair and Replacement  
COMPUTER ASSEMBLY  
When it is determined that a problem exists in the computer assembly itself,  
the entire assembly must be sent back for repair. To remove the computer  
assembly, proceed as follows:  
1) Ensure that the main power cord is  
unplugged from the wall outlet.  
2) Remove the two side panel  
mounting screws from the console as  
shown to the right.  
3) Lift up the side panel and remove it.  
4) Remove the two screws from the  
front edge of the inner panel and swing  
it out to expose the computer  
assembly.  
5) Carefully remove all of the cables from the rear of the computer assembly  
noting where each individual cable was connected.  
6) The computer assembly is held in place by four bolt assemblies with rubber  
bumpers on them. Relieve the tension on all four bolt assemblies until the  
computer moves easily forward and backward.  
7) Remove the computer assembly from the front of the tester and send it to  
Digalog for repair.  
To reinstall the computer assembly, reverse this procedure.  
Series 2040 Test System  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Module Repair and Replacement  
Series 2040 Maintenance Manual V2.00  
INTERNAL AND PATCHBOARD POWER SUPPLY OVERVIEW  
The Testhead Power Supply Controller monitors Testhead power supplies for  
overvoltage with OVP crowbar IC’s. When an overvoltage is detected, the  
OVP IC quickly short circuits the offending power supply. This produces an  
undervoltage state on that power supply. This is, in turn, sensed by an opto-  
isolated comparator to produce a digital fail signal. This digital signal is latched  
in the control circuit and removes the AC line power from the Testhead  
Internal and/or Patchboard Power Supplies. The state of all comparators is  
indicated by the LEDs mounted on the top edge of the controller circuit card.  
An illuminated LED in the power supply group indicates a normal condition.  
The “PBON”, “RESET”, “PROX” and “EXTERNAL FAULT” LEDs, when all are  
illuminated, indicate that power is applied to the patchboard. The  
extinguishing of any LED shows the reason for the loss of power to the  
Patchboard and/or the Testhead. A dedicated power supply on the controller  
circuit card maintains its function without regard to the monitored supplies.  
The result of the logical OR of any Testhead Internal Power Supply failure is  
logically OR’d with any Patchboard Power Supply failure. The result of this is  
provided as a current loop signal through an opto-isolator to the UUT Power  
Supply Controller. Therefore, when a fault is detected in the Testhead Internal  
Power Supply, all power supplies are shut down. If the fault originates in the  
Patchboard Power Supply, the Testhead Internal Power Supply remains on.  
The Testhead Power Supply Controller also responds to three external logic  
signals. They are the Patchboard Open Sensor, CPU ON/OFF command and  
Reset. A proximity detector, located near the Patchboard, senses the release of  
the Patchboard and causes a shutdown of the Patchboard Power Supplies,  
and indirectly the UUT Power Supplies. The CPU ON/OFF command (see the  
functional call POWER) is a programmed command where a “0" causes the  
Patchboard Power Supplies to shut down without a detected fault condition.  
A ”1" enables the Patchboard Power supplies. The Patchboard Power is  
removed when a Testhead reset is issued. (See the TCLEAR functional call in  
the Programming Manual). When the Patchboard Power Supplies are shut  
down, they are also disconnected from the Patchboard via relays.  
Series 2040 Test System  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Module Repair and Replacement  
TESTHEAD & PATCHBOARD POWER SUPPLY REPLACEMENT  
WARNING! Remove main power from tester before proceeding.  
1) Ensure that the main power cord is  
unplugged from the wall outlet.  
2) Remove the two side panel  
mounting screws from the console as  
shown to the right.  
3) Lift up on the side panel and  
remove it.  
4) Remove the two screws from the  
front edge of the inner panel and swing  
it out to reveal the four power supplies.  
5) Locate the proper Testhead or Patchboard power supply connector for the  
supply to be replaced and disconnect it from the Testhead Power Supply  
Controller. Use the legend below.  
6) Remove and replace the  
faulty power supply, taking  
care to route the new cable  
exactly the same as the old  
one.  
J7 - +5VDC, +15VDC  
J8 - -5.2VDC, +15VDC  
J9 - +19VDC  
J13 - +5VDC  
Series 2040 Test System  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Module Repair and Replacement  
Series 2040 Maintenance Manual V2.00  
+5VDC Testhead Power Supply  
The Testhead cooling fan assembly contains the main +5VDC Power Supply  
for the Testhead. If it is determined that the supply is faulty, the recommended  
procedure would be replacement of the entire assembly as described below:  
1) Remove the four screws on the sides of the Testhead cooling fan assembly,  
and draw the assembly towards you 2 to 3 inches.  
X
Y
Z
1
0
M
HZ  
E
XT  
C
LK  
E
XTEV N T  
E
G
PIB  
P
RIN T  
E
R
ET  
HERNE  
T
C
O
M
1
KE  
Y
B
O
A
RD  
M
O
DEM  
Fasteners  
2) Remove the five fasteners for the filter housing plate as shown above, and  
remove the plate from the cooling fan assembly.  
3) Loosen the two Testhead retaining screws located along the back of the  
black Testhead cover, by turning them 1/4 turn counterclockwise. Using the  
handle, lift the Testhead into the position shown on the next page.  
4) Remove the hazard cover from the Testhead by removing the mounting  
screw from each end of the cover.  
5) Disconnect the supply wiring harness from connector J13 on the Testhead  
Power Supply Controller board.  
Series 2040 Test System  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Module Repair and Replacement  
Testhead  
Boards  
Retaining  
Screws  
Remove the assembly  
through this opening.  
6) Disconnect the AC fan power cord from the Testhead section of the AC  
Distribution box.  
7) Remove and replace the assembly through the Testhead bay, taking care to  
route the wiring for the new Testhead fan assembly exactly as the old wiring  
was routed.  
8) To reinstall the supply, reverse steps 1 - 7.  
Series 2040 Test System  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Module Repair and Replacement  
Series 2040 Maintenance Manual V2.00  
UUT POWER SUPPLIES - Description  
Contained in the middle rack of the unit are the product power supply  
controllers. Each one controls a single product supply. The controllers are  
arranged PCS2, PCS1, and PCS0 from left to right. (This is a typical installation.  
Individual systems may vary.)  
To the right of the controllers, PS2, a Kikusui PDA110-1.5L, completes the top  
power supply rack. Directly beneath, a second rack contains, from left to right,  
PS1, a Kikusui (or Kenwood) PD-35-20, and a Kikusui PD55-10 (Kenwood  
PD56-10). The bottom rack is heavy, so care is suggested in handling it.  
The three power supply controllers are identical. To adapt them to the  
different characteristics of each supply, a configuration card, unique for each  
supply, is inserted into the front of each controller. To replace a configuration  
card, ensure that the power to the controller is shut off, and pull the black tabs  
and remove the card. To install, reverse the process.  
Series 2040 Test System  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Module Repair and Replacement  
UUT POWER SUPPLY CONTROLLER REPLACEMENT  
WARNING! Remove main power from tester before proceeding.  
To remove a controller, remove the upper and lower retaining screws from the  
front of the console. The controller should slide directly forward. Place the  
controller to be installed on the floor directly beside the controller being  
replaced. Place it in the same orientation to make the process simpler. One by  
one, remove each of the cables/cords from the back of the controller being  
removed, installing them in the exact same locations on the replacement  
controller. To reinstall, reverse the removal process.  
Series 2040 Test System  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Module Repair and Replacement  
Series 2040 Maintenance Manual V2.00  
UUT POWER SUPPLIES  
WARNING! Remove main power from tester before proceeding.  
The removal process for UUT Power Supplies and GPIB/HPIB Power Supplies  
is almost identical. However for both types of supplies, extreme caution  
should be exercised due to the considerable weight of the supplies.  
Lower Rack Power Supplies (Kikusui or Kenwood PD-35-20, Kikusui PD55-  
10, and Kenwood PD56-10)  
1) Remove the left side panel (side opposite the computer) to gain access to  
the rear of the power supplies.  
2) Disconnect the AC power cords from the AC/DC Distribution Box, and  
remove the Power Output/Programming cables from the UUT Controllers.  
3) Remove the screws in the front panel holding the lower rack in place, and  
slide the entire rack forward out of the cabinet. The rack assembly is very  
heavy so proceed with caution.  
4) Remove the faulty supply from the rack and return for repair.  
Upper Power Supplies (PDA110-1.5L)  
Repeat the above procedure except that the supply can just be removed by  
removing the small upper and lower retaining brackets. No rack assembly  
needs to be removed.  
Series 2040 Test System  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Module Repair and Replacement  
GPIB/HPIB POWER SUPPLIES  
1) Remove the left side panel (side opposite the computer) to gain access to  
the rear of the power supplies.  
2) Disconnect the AC power cords from the AC/DC Distribution Box, and  
disconnect the GPIB cable, Fault/Inhibit cable, and power output cable from  
the rear of the supply.  
3) If the faulty supply is the lower of two GPIB/HPIB supplies, it may be easier  
to remove the upper supply first.  
4) Remove the faulty supply and return for repair.  
GPIB/HPIB  
Supplies  
PAD115-1.5L  
Kikusui PD-35-20,  
Kenwood PD-35-20,  
Kikusui PD55-10, or  
Kenwood PD56-10  
This illustration shows a typical installation. Individual systems may vary from  
this example.  
Series 2040 Test System  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Module Repair and Replacement  
Series 2040 Maintenance Manual V2.00  
UUT - GPIB/HPIB POWER SUPPLY DISCONNECT AT TEST BAY  
From top to bottom:  
PS0  
PS1  
PS2  
PS3 - Jumper  
PS4 - Jumper  
All unused power supplies must be jumpered. This is a typical installation.  
Individual systems may vary. Also note, the GPIB/HPIB power supplies may be  
externally connected to the fixture/UUT.  
Series 2040 Test System  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Module Repair and Replacement  
CIRCUIT BOARD REPLACEMENT  
The illustration below shows the positions of the boards in a typical Series  
2040 Test System. The MDE, TMS, and AMS are always the last three on the  
left side of the tester (as viewed from the rear of the Testhead) and are  
dedicated to those slots shown. The PSD (Power Supply Distribution)  
assembly always occupies Slot 0, while the PSC (P/S Controller) occupies its  
own slot on the far right.  
26  
25  
24  
23 22 21 20 19 18 17 16 15 14 13 12 11 10 09 08 07 06 05 04 03 02 01 00  
To remove a circuit board, grasp the white tabs on the corners of the board  
and spread them outward towards the sides of the Testhead, as shown on the  
next page.  
Series 2040 Test System  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Module Repair and Replacement  
Series 2040 Maintenance Manual V2.00  
Spread these tabs.  
Spread these tabs.  
Pin Guide  
To reinstall a board, first check the positioning of the black plastic pin guide to  
ensure that the beveled openings are aligned with the receptacles on the  
board. Position the board’s leading edge in the top and bottom card guides for  
the Testhead slot and carefully insert the board until resistance is felt. At this  
time, it is crucial to ensure that the pointed ends of the Patchboard pins are  
properly aligned with the receptacles on the board. When the alignment is  
correct, a slight “give”will be felt as the pins begin to seat. At this point, place  
your thumbs on the white tabs and firmly seat the board. Ensure that it is fully  
seated by comparing the position of the rear edge of the board with the rear  
edge of neighboring boards.  
If excessive resistance is encountered or a high pitched metallic noise is heard,  
stop and remove the board or damage to  
the pins and/or receptacles may occur.  
Repeat the procedure from the beginning.  
In some cases, temporary removal of boards  
in adjacent slots may be helpful.  
Series 2040 Test System  
62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Module Repair and Replacement  
CONTACT SPRING REPAIR  
Alignment of the contact springs (pins) in the Patchboard receiver of the  
Digalog Series 2040 Test System is critical. Misaligned or bent pins can cause  
failures in the system. It is possible that one or more contact springs may  
become damaged in one way or another. If the springs are bent, and the bend  
is horizontal (lateral), the springs may be repairable. If the springs are severely  
bent, or bent vertically, they must be replaced.  
When damage to the springs is minor and  
repairable, they may be straightened with  
an Alignment Comb (Digalog part number  
0000-2745) designed for this purpose.  
When several contact springs of a single  
row are out of position, straighten them  
with the alignment comb to attain  
uniformity. Use the comb as follows:  
1. De-energize the system, and remove the Patchboard fixture to expose the  
contact springs.  
2. Determine the contact springs to be straightened. You can do this by  
placing a non-metallic straight edge vertically along the row to be straightened.  
CAUTION: Be sure the contact springs you are aligning are either all left  
hand or all right hand springs.  
3. Place the bottom edge of the teeth of the  
comb between two vertical rows of contact  
springs.  
4. Place the bottom edge of the teeth against  
the base of the contact springs in the row  
adjacent to those being straightened.  
5. Push against the top of the contact springs  
with the comb. It may be necessary to bend  
the contact springs beyond the normal  
Series 2040 Test System  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Module Repair and Replacement  
Series 2040 Maintenance Manual V2.00  
position to ensure that they all have the proper spring tension. If so, you must  
then place the comb on the opposite side of the contact springs and bend  
them back to the normal position.  
Individual contact springs can be aligned using a suitable non-metallic tool  
(e.g. the eraser end of a pencil). Simply reposition the bent contact spring so  
that it is aligned with the properly positioned springs in the same row.  
Alignment of the contact springs can be visually checked by looking straight  
through the holes of an empty Patchboard with the Patchboard in operating  
position. When making such a visual check, be sure that the contact springs  
are positioned within the range shown.  
If a contact spring is severely bent, or bent vertically, replacement of the  
contact spring is recommended. Should the bend be minor and in the  
horizontal (lateral) plane, the contact spring may be repairable.  
USE OF PATCHBOARD SPRING REMOVAL TOOL (0000-2746)  
Determine the slot in which the damaged spring is located. If there is a  
Testhead card in that slot, it will need to be removed. Be sure to use a wrist  
strap when removing or inserting the card, and be sure to place the card into a  
static bag while it is out of the tester.  
Turn off Testhead power and open the Testhead. Remove the card if  
necessary, put it into a static bag and set it safely aside. Determine the  
location of the damaged spring. Observe on the top (fixture mating) side of  
the Testhead that each spring has a pair of small metal tangs at the base,  
opposite the “finger” of the spring. The latching tab on the bottom (Testhead  
card) side is on the same side of the spring as the tangs.  
Series 2040 Test System  
64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series 2040 Maintenance Manual V2.00  
Module Repair and Replacement  
Retract the plunger of the spring removal tool and position the sleeve of the  
tool so that the flat side of the “D”-shape formed by the end of the sleeve is  
aligned with the latching tab of the damaged Patchboard spring (Testhead card  
Board Side of Patchboard  
“D”shaped  
latching tab  
“D” shaped sleeve  
Plunger  
Patchboard  
side of the Patchboard). Carefully lower the sleeve over the spring until it is  
fully seated and the sleeve is resting against the Patchboard. The “stinger” of  
the spring will slide into the hollow end of the plunger. A flashlight may be  
needed to provide adequate lighting, especially if there are Testhead cards on  
either side of the slot. The latching tab of the spring should be fully depressed  
inside the tool sleeve.  
Be sure that the tool is held straight in-line with the spring. Apply steady,  
careful pressure on the handle. Do NOT push sharply and do NOT strike or  
tap the end of the handle. Resistance will be felt as the spring slides out of the  
Patchboard. Continue to apply steady pressure until the plunger handle  
bottoms out. If these precautions are ignored, it is possible to damage the tool  
or the Patchboard material itself.  
The ejected Patchboard spring will be protruding from the top (fixture mating)  
side of the Patchboard. Simply grasp it and pull it free of the Patchboard.  
After replacing the spring in the Patchboard, re-install the Testhead card and  
close the Testhead.  
CONTACT SPRING REPLACEMENT (Conventional Method)  
To remove a damaged contact spring when the removal tool is not available,  
use the following procedure:  
Series 2040 Test System  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Module Repair and Replacement  
Series 2040 Maintenance Manual V2.00  
1. Remove the termination from the rear of the spring. (Remove the Testhead  
card.)  
2. With a slow and steady pressure, bend the one side of the “U” in far  
enough to clear the board opening. Then, in the same manner, bend the other  
side of the “U” in far enough to permit the contact spring to pass freely though  
the board cavity.  
CAUTION: To prevent chipping of the cavity, double check that the “U” will  
clear the cavity opening.  
3. Face the front of the board, and grasp the damaged contact spring across  
the chevron with a needle-nosed pliers. With a steady, even pressure, pull  
straight out on the contact spring until it is free and clear of the board - and  
then destroy the contact spring immediately!  
To install a new contact spring in the board, the following steps are  
recommended:  
1. Be sure that you have the proper contact spring, right hand or left hand,  
whichever is required (the chevron contact area should be facing the general  
direction of the bottom left or bottom right corner of the frame).  
2. The Digalog part numbers for the springs are as follows:  
Right Hand: 2825-1011  
Left Hand:  
2825-1012  
3. Orient and start the contact spring by hand, according to the position of the  
“D” within the board cavity.  
4. Place the end of a medium size screwdriver against the end of the barrel, at  
the base of the blade. Then press with a slow and even pressure until the  
“ears” are seated on the board. As the “ears” reach the surface of the board,  
you will feel the locking tab snap into the locked position.  
5. Check the replacement contact spring to see that it is properly aligned with  
the adjacent springs.  
Series 2040 Test System  
66  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Haier Washer HLP021 WM User Manual
Hamilton Beach Blender All Metal Blender User Manual
HANNspree Flat Panel Television 96 Liquid Crystal Display User Manual
Healthrider Home Gym HREL11900 User Manual
Heath Zenith Work Light SL 5650 User Manual
Hotpoint Freezer RFA60 User Manual
Hotpoint Refrigerator FF50 User Manual
HP Hewlett Packard Computer Accessories 16047C User Manual
IBM Blood Glucose Meter 5697 VM3 User Manual
IBM Laptop Z50 User Manual