Bacharach Smoke Alarm ADM800 User Manual

AGM300 / ADM800  
Ammonia Gas Monitoring System  
Instruction 3015-4275  
Installation / Operation / Maintenance  
Rev. 3 – May 2006  
INSPECTION EQUIPMENT  
MEASURING EQUIPMENT  
61KN  
Product Leadership • Training • Service • Reliability  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Table of Contents  
Instruction 3015-4275  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
How to Use This Manual  
Thank you for investing in a BACHARACH Ammonia Gas Monitoring System.  
This manual provides important information on how to install, operate, and service the AGM300 Ammonia Gas  
Monitor and ADM800 Ammonia Display Module.  
Please Read this Manual Carefully Before Use  
If you have a working knowledge of gas monitors, you will find this manual useful as a reference tool. If you are  
new to the use of gas monitors, you can educate yourself about the principles of gas detection and the proper  
operation of this device by reading this manual thoroughly.  
ii  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Safety Precautions  
AC Power Supply  
Ensure the source voltage matches the voltage of the product before energizing the equipment. It is highly  
suggested that the AGM300 be placed on a separate circuit (with UPS or surge protection).  
A switch or circuit-breaker shall be included in the building installation  
It shall be in close proximity to the equipment and within easy reach of the operator  
It shall be marked as the disconnecting device for the equipment  
Protective Grounding  
Under no circumstances should this equipment be operated without connection to a protective ground. Doing  
so poses a potential shock hazard and is also a violation of electrical safety standards applicable to this type  
of equipment.  
Explosive Atmosphere  
Do not operate this equipment in the presence of flammable liquids, vapors or aerosols. Operation of any  
electrical instrument in such an environment constitutes a safety hazard.  
Proper Exhaust Venting  
It is imperative that the exhaust port on this instrument be properly vented as described in this manual. Failure  
to do so constitutes a safety hazard.  
Working Inside the Instrument  
Extreme care should be exercised when accessing the interior of this instrument. Only qualified electrical  
maintenance personnel should perform connections and adjustments. Always de-energize the power supply  
before working inside the instrument.  
Modifications to the Instrument  
Under no circumstances should this instrument be modified without written consent from Bacharach, Inc.  
Changes or modifications to this instrument, not expressly approved, will void the warranty.  
In Case of Malfunction  
Do not continue to use this equipment if there are any symptoms of malfunction or failure. In the case of such  
occurrence, de-energize the power supply and contact a qualified repair technician or the nearest Bacharach  
Service Center. ONLY provided knockouts are to be used for electrical and communication wiring. Drilling into  
the box will void the warranty.  
Fusing  
AGM300, 120 VAC Input, Power Supply P/N 3015-3167  
F1: 0.5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2538  
F2: 250 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2537  
F3: 5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2539  
F4: 63 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2535  
AGM300, 230 VAC Input, Power Supply P/N 3015-3161  
F1: 250 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2537  
F2: 125 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2536  
F3: 5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2539  
F4: 32 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2534  
ADM800  
F1 & F2: 1 A, 250VAC, Fast Acting, 5mm x 20mm, P/N 3015-4011  
Instruction 3015-4275  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Safety Precautions (Cont.)  
Installation Category  
Installation Category II, Pollution Degree II, as defined by UL.  
Altitude Limit  
2,000 meters  
Cleaning  
USE a dry cloth to clean the outside of the case. DO NOT use soap and water.  
Warning Statements  
The use of the word WARNING in this manual denotes a potential hazard associated with the use of this  
equipment. It calls attention to a procedure, practice, or condition, or the like, which if not correctly performed or  
adhered to, could result in personal injury or death.  
Caution Statements  
The use of the word CAUTION in this manual denotes a potential hazard associated with the use of this  
equipment. It calls attention to a procedure, practice, condition, or the like, which if not correctly performed or  
adhered to, could result in damage to the equipment.  
Hazard Symbols on Monitor  
This symbol indicates the need to consult this operating instruction  
manual when opening the enclosure.  
WARNING: A potential risk exists if the operating instructions are not  
followed.  
This symbol indicates the presence of electric shock hazards when  
the enclosure is opened.  
WARNING: To avoid risk of injury from electric shock, do not open  
the enclosure when power is applied.  
iv  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Functional Overview  
General Description  
The AGM300 is designed to support compliance to federal, state and local safety codes governing ammonia  
gas emissions. Avoiding significant ammonia gas loss reduces equipment replacement costs, maintains  
equipment efficiency, promotes safety, and protects the environment.  
The AGM300 provides for the continuous monitoring of ammonia gas levels in up to 16 separate test zones.  
Each zone can be independently programmed to identify leak (small), spill (medium), or evacuation (large)  
levels of gas. The instrument also retains a log of previous readings that can be easily accessed for analysis.  
An audible alarm and large LED indicators are provided to signal alarm conditions, and relay contacts are  
provided for the connection of external alarm devices. The system also includes two 4–20 mA current loop  
interfaces (optional) for connection to external devices.  
The AGM300 requires only minor periodic maintenance such as the occasional replacement of filters. The  
instrument incorporates active diagnostics that continuously monitor the system for proper operation. An LED  
indicator is provided to indicate system malfunctions, and fault codes are displayed that enable the operator to  
identify the source of the fault.  
Communication Options  
The AGM300 features full two-way communications via an RS-485 interface. MODBUS RTU is the  
communication protocol standard. The instrument can be connected directly to a Building Management System  
or it may be operated as a stand-alone system using the ADM800 Ammonia Display Module.  
An RS-232C port is also provided for connection to a PC. This enables the AGM300 to be setup from a  
personal computer.  
Please refer to the Appendix for a more complete discussion of communication protocols.  
Suggested Location of Sampling Points  
At the point of an ammonia gas leak the gas is nearly pure. As the gas is dispersed into the air, the gas  
molecules diffuse causing a dilution of the original concentration. The AGM300 measures the ammonia gas  
concentration at the precise point the sample is collected. Therefore, if the termination of the collection line is  
not at the exact point of the gas leak, then the unit will read a diluted mixture of the gas and air.  
It should also be noted that when ammonia gas is cold it is heaver than air and settles below the leak point, but  
as the gas warms to room temperature it becomes lighter than air and tends to collect above the point of a leak.  
Consequently, sampling points should ideally be located as close as possible to the source of potential leaks. If  
this is impractical, then the alarm set points should be adjusted for that zone to compensate for the dilution of  
the gas. DO NOT plug any of the zones. Plugging a zone will give the monitor a false indication during  
start up.  
The AGM300 should be centrally located in the mechanical room and be readily accessible for easy visual  
monitoring and servicing. Air sample tubing may be run in lengths up to 500 feet. The fresh air purge line  
should draw from an area that does not contain any ammonia gas and cannot exceed 300 feet in length. The  
exhaust line should run to an out side location if possible. The length of the exhaust line cannot exceed  
300 feet.  
Ideally, two to three pick up points spaced around each chiller will provide sufficient coverage. It may be  
necessary to perform a “smoke” test of the mechanical room to determine the best locations. The smoke test  
would provide the pattern of air currents present in the mechanical room.  
Instruction 3015-4275  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AGM300 / ADM800 Refrigerant Gas Leak Monitor Mechanical Room Placement  
Aux. Horn/Strobe  
Remote Alarm  
Fresh air purge  
from area away  
from ammonia  
gas  
Exhaust Fan  
Outside Hallway  
Sample Inlet  
Pickup Point  
ADM800  
Ammonia Display  
Module outside of  
mechanical room  
Machine Room  
AGM300  
Ammonia  
Gas  
Monitor  
Chiller  
Sample Inlet Pickup Points  
The ADM800 should be mounted outside of the mechanical room or at least just inside of a door to the room.  
This is the “split architecture design” for safety of the operator. The ADM800 can be located up to 4500 feet  
from the AGM300. The ADM800 is the main interface by which you program the AGM300, acknowledge  
alarms, and observe conditions inside of the mechanical room. Note that there are two additional alarm relay  
contacts in the ADM800 that can be programmed to alarm with “leak, spill, evacuate, fault or monitor on”.  
vi  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
AGM300 Installation  
Upper Mounting  
Bracket  
LED  
Indicators  
Lower Mounting  
Bracket  
Water Trap  
STANDARD ACCESSORIES FOR A 4 POINT SYSTEM  
5 Line-End Filters (P/N 3015-3420)  
AGM300/ADM800 Instruction Manual (P/N 3015-4275)  
Instruction 3015-4275  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AGM300 - Installation Considerations  
Locating the Monitor  
The AGM300 should be centrally located in the facility and should be easily accessible for visual monitoring and  
servicing. Intake sample lines can be up to 500 feet in length, but it is important to remember that sampling  
cycle time is proportional to the total number and length of individual sample lines.  
Dirt, grease, and oils can adversely affect the operation of the AGM300. The monitor should be installed out of  
direct sunlight in a clean, dry area that is not subject to temperature or humidity extremes. Installation of the  
monitor in a mechanical room is acceptable provided reasonable environmental conditions exist. If there is a  
question, consider installing the unit outside of the mechanical room in a cleaner area of the facility.  
Warnings and Cautions  
WARNING Drilling holes in the AGM300 and  
ADM800 enclosure will void the warranty. Please  
use knockouts provided for electrical connections.  
WARNING - Do not operate this equipment in the  
presence of flammable liquids, vapors or aerosols.  
Operation of any electrical instrument in such an  
environment constitutes a safety hazard.  
WARNING - Always de-energize the power supply  
before working inside the instrument.  
CAUTION - The AGM300 contains sensitive  
electronic components that can be easily damaged.  
Be careful not to touch or disturb any of these  
components.  
Inspection  
The AGM300 has been thoroughly inspected and tested prior to shipment from the factory. Nevertheless, it is  
recommended that the monitor be re-checked prior to installation. Inspect the outside of the enclosure to make  
sure there are no obvious signs of shipping damage. Open the door latches and inspect the interior of the  
instrument for loose components that may have become dislodged during shipment. If damage is discovered,  
please contact the nearest Bacharach Service Center for assistance.  
2
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AGM300 - Mounting Instructions  
Screw Locations  
The AGM300 should be installed plumb and level and securely fastened to a rigid mounting surface. The  
enclosure utilizes keyhole mounting brackets designed for #12 pan head fasteners. Locate the four screws as  
shown in the diagram and leave the screw heads protruding approximately 3/16".  
AGM300 Mounting Specs  
Installation  
Hold the monitor flat against the mounting surface and allow it to slide down engaging the screw heads in the  
keyhole slots of the mounting brackets. Adjust the screws as necessary to hold the monitor securely against the  
mounting surface.  
Instruction 3015-4275  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AGM300 - Connecting Air Lines  
Overview  
Individual sample lines are run from the AGM300 to each area of the facility to be monitored. Additionally, a  
purge line is installed to provide clean air for resetting the infrared zero baseline. An exhaust line is installed to  
vent residual gas away from the monitor. All sample line connections are made on the left side of the enclosure  
as indicated in the photograph below.  
AGM300 Side View  
Intake Ports  
Exhaust Port  
Purge Port  
RS-232 Connector  
RS-232  
Inlet/Purge Ports  
(see below)  
Exhaust Port  
Inlet / Purge Ports  
4
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Tubing Considerations  
Use ¼" outside diameter (0.040" wall) flex tubing is used for all air lines (P/N 304-2742). The tubing should be  
clean and free of residual moisture or other contaminants. The tubing should be cut cleanly with a sharp knife  
and care should be taken not to distort the tubing end.  
IMPORTANT: Due to the reactive nature of ammonia gas, it is important to use materials that will not absorb  
ammonia gas as it passes though the sampling system. The use of unauthorized tubing or any other materials  
in the gas sampling stream may lead to erroneously low readings.  
To connect the air lines to the monitor simply push the tubing firmly onto the connector. All tubing bends should  
have a radius of no less than 5" to assure proper airflow. If kinks or obstructions occur in any of the air lines the  
instrument may not function properly.  
Connecting Purge Line  
A purge line is required to draw fresh air into the instrument and should not exceed 300 feet in length. It is  
advisable to terminate the line outdoors, provided the input is not exposed to rain, snow, ice, exhaust fumes, or  
other airborne contaminates. If an outdoor installation is impractical, the line should be run to an area inside the  
facility that you are certain is not contaminated with ambient ammonia gas. A line-end filter (P/N 3015-3420)  
should be attached to the end of the purge line.  
Connecting Exhaust Line  
An exhaust line is required to vent gas samples away from the instrument and should not exceed 300 feet in  
length. The exhaust line should terminate in a location that is completely isolated from the purge line  
termination point and other areas of the facility that will be monitored. Ideally this line should terminate outdoors  
in a location that is not exposed to the elements. This line does not require a line-end filter. If the exhaust line  
terminates outside the building, position the tubing so that no water or moisture can enter it.  
Connecting Sample Intake Lines  
The AGM300 is designed to accommodate up to 16 separate sample intake lines. The standard configuration of  
the unit includes one manifold of 4 intake connectors and 1 purge connector. Additional manifolds can be  
easily installed to increase monitoring capacity (field installation kit part P/N 3015-3419, and 4 zone line end  
filter kit P/N 3015-3411).  
Sample intake lines can be up to 500 feet in length. All line terminations should be positioned to reduce the  
possibility of mists, aerosols, oil, water, dust, or other contaminates being drawn into the instrument. A line-end  
filter (P/N 3015-3420) should be attached to the end of each sample intake line.  
IMPORTANT: DO NOT plug any of the zones. Plugging a zone will give the monitor a false indication during  
start up.  
Please refer to the earlier Section Suggested Location of Sampling Points to learn more about where to place  
the ends of the sample intake lines.  
Water Trap  
A water trap has been installed as a standard feature. The water trap prevents condensation or moisture from  
entering the infrared device and causing serious damage. To empty the water trap, loosen the wing nut and  
allow water to drain – be sure to retighten the wing nut.  
CAUTION - The introduction of contaminants through the  
air intake lines can result in serious and permanent  
damage to the monitor.  
Instruction 3015-4275  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
AGM300 - Interior Schematic  
Primary Filter  
Intake Manifold Solenoids  
Intake Solenoid Connectors  
Power Supply Board  
Main Board  
Microcontroller Board  
Infrared Optical Bench  
Gas Sample Pump  
Primary Filter (P/N 3015-3071)  
Power Supply  
Board  
120 or 230 VAC  
Determined by  
Sales Order  
Intake  
Manifold  
Solenoids  
Main Board  
Intake  
Manifold  
Solenoid  
Connectors  
Microcontroller  
Board  
Infrared  
Optical Bench  
Gas Sample  
Pump  
6
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AC Power / RS-485 / Relay Connector / 4-20 mAdc Current Loop  
AC Power  
ON/OFF  
Switch  
4–20 mAdc  
Interface Board  
(Optional)  
AC IN  
Connector  
T1  
Dual 4-20 mAdc  
Outputs  
(Signal Out Only)  
DO NOT APPLY  
POWER!  
Location of  
AC Input  
Line Filter  
RS-485  
Connector  
(230 VAC  
Units Only)  
Relay Connector  
Ground Stud  
Auxiliary AC Out Connector T2  
Alarm 1, 2, 3, Fault  
(Used to Supply AC power to Alarm Devices)  
(Overload Protected by Fuse F3)  
Switches / Fuses  
F4  
F2  
F1  
AGM300  
Fuses  
(Refer to  
Maintenance  
Section for  
values Page 64)  
F3  
Connects to Aux. AC Out  
Connector T2  
5 Amp 250VAC  
(P/N 604-2539)  
AC Power  
ON/OFF  
Switch  
Factory  
Default  
Terminator  
IN / OUT  
CPU  
Reset  
Node Address  
Switches  
Instruction 3015-4275  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AGM300 - Electrical Wiring  
It is highly recommended that the AGM300 be connected directly to the AC power source, preferably on its own  
circuit. The connection should be completed with a UL rated multi-conductor wire (14-18 AWG).  
Locate service knockouts on exterior of enclosure and install electrical conduit in the typical manner.  
The AGM300 is available in either a 120 or 230 VAC configuration. The unit’s AC voltage rating is marked on  
the outside of the enclosure. Do one of the following to connect the AC power leads and ground:  
• 120 VAC Installation  
Locate the AC IN connector T1 and the Ground Stud on the inside of the enclosure (Page 7).  
Remove the AC IN connector from the Power Supply board, and then secure the incoming AC power line  
(black) and neutral (white) wires to this connector as shown in the following illustration. Carefully plug the  
connector back onto the circuit board.  
Using the supplied crimp-on ring terminal, washers, and nuts, connect the AC input ground wire (green) along  
with the ground wire connected to the enclosure’s lid to the monitor’s Ground Stud.  
• 230 VAC Installation  
Locate the AC Input Line Filter’s black and white wires, and the Ground Stud on the inside of the enclosure  
(Page 7).  
Using wire nuts, secure the incoming AC power Line 1 (black) and Line 2 (white) wires to the AC Input Line  
Filter as shown in the following illustration.  
Using the supplied crimp-on ring terminal, washers, and nuts, connect the AC input ground wire (green) along  
with the ground wire connected to the enclosure’s lid to the monitor’s Ground Stud.  
WARNING – Drilling holes in the AGM300 enclosure will  
void the warranty. Please use knockouts provided for  
electrical connections.  
WARNING - Electrical installation should be performed  
by a certified electrician and should comply with all  
applicable local, state, and federal electrical safety  
codes.  
WARNING - Under no circumstances should this  
instrument be operated without connection to a  
protective ground. Doing so poses a potential shock  
hazard and is also a violation of electrical safety  
standards applicable to this type of equipment.  
A switch or circuit-breaker shall be included in the building installation  
It shall be in close proximity to the equipment and within easy reach of the operator  
It shall be marked as the disconnecting device for the equipment  
8
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AGM300 AC Input Power and Ground Connections  
Instruction 3015-4275  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
AGM300 - Connecting Communication Devices  
Ammonia Display Module (ADM800) Connection  
The AGM300 is connected to the ADM800 using a shielded twisted pair instrument cable. The maximum  
distance between the farthest away AGM300 and ADM800 is 4500 feet.  
Use any of the remaining service knockouts to gain access to the interior of the monitor. The RS-485  
communication wiring between the AGM300 and ADM800 must be connected in the following manner:  
Locate the RS-485 connector in the AGM300 (Page 7). Connect one lead of a twisted shielded pair to the  
“B” connection point (the far left point), note the wire color. Connect the second wire to the “A” connection  
point (the middle), note the wire color. Connect the ground to the “GND” connection point  
Locate the RS-485 connector marked “TO MONITORS” in the ADM800 (Page 23). This connector is  
located on the far-left bottom of the ADM800 PC board. Make the wire run to the ADM800 and connect the  
twisted shielded pair to the RS-485 TO MONITORS connector using the same color code as used on the  
AGM300.  
Integrating with Building Management Systems  
The AGM300 may be connected directly to a Building Management System using a shielded twisted pair cable.  
The cable from the Building Management System is connected to the RS-485 connector inside the AGM300.  
MODBUS RTU is the standard communication protocol.  
Use any of the remaining service knockouts to gain access to the interior of the monitor. Locate the RS-485  
connector and remove it from the circuit board. Secure the wire leads to the connector orienting them as shown  
in the diagram below. Check to make sure that the polarity matches the wiring to the Building Management  
System. When you are through securing the connections, carefully plug the connector back onto the circuit  
board.  
Larger Integrated Systems  
You may also connect the AGM300 to a Building Management System through an ADM800. In this case, first  
connect the AGM300 to the ADM800 as described above. Then, follow the instructions under the heading  
ADM800 – Communications Connections on Page 26 for information on how to connect the ADM800 to a  
Building Management System.  
RS-485 Connector  
10  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Changing Terminator Settings  
The terminator switch is shipped from the factory in the terminated or “IN” position. This is the correct setting if  
the AGM300 is connected as a single device, or it is the last device on the network chain. If the AGM300 is  
being installed in the middle of a network, the terminator must be moved to the “OUT” position.  
Locate the switch and determine its position. If it needs to be moved, slide the switch to the appropriate  
position.  
Network Node  
Address  
Switch  
Terminator  
Switch  
IN  
OUT  
Node Address DIP Switch Settings  
(Refer to Multiple AGM's and Connecting to a Building Management System)  
Each AGM300 on the network must have a distinct node address. The node address may be set from 1 to 15.  
Four dip switches numbered 1, 2, 4, 8 on the main circuit board are used to define this value by setting the  
correct combination of dip switches to their ON positions. The node address is the sum of the switch numbers  
that are ON.  
Example: - For node address “5” switches 1 and 4 are ON.  
- For node address “13” switches 1, 4, & 8 are ON.  
Power must be cycled to elect this change.  
Personal Computer  
(Refer to AGM300 PC Software Section)  
The AGM300 may be connected to a personal computer using the RS-232 interface on the left side of the  
enclosure.  
Software will be provided upon request or as a download from:  
Instruction 3015-4275  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Multiple AGM’s  
TERMINATOR  
“IN”  
MULTIPLE AGM300’S WITH ADM800  
NOTE 1: The last AGM300 or ADM800 on either end of the network must have its terminator  
in the “IN” position, and all other units must have their terminators in the “OUT” position.  
AMD800  
NOTE 2: The total length of the RS-485 cable cannot exceed 4500 feet. (Use instrument  
cable 20 gage multi-strand shielded and twisted pair – similar or equal to Belden  
cable #8762.)  
RS-485 CONNECTIONS BETWEEN AGM300’s  
NOTE: The terminators in the ADM800 and AGM300 Unit 4 must be in the “IN” position. The  
terminators in AGM300 Units 1, 2 & 3 must be in the “OUT” position.  
RS-485  
AGM300  
UNIT 1  
AGM300  
UNIT 2  
AGM300  
UNIT 3  
AGM300  
UNIT 4  
NODE 1  
NODE 2  
NODE 3  
NODE 4  
TERMINATOR  
“OUT”  
TERMINATOR  
“OUT”  
TERMINATOR  
“OUT”  
TERMINATOR  
“IN”  
Connecting to a Building Management System  
The AGM300 may be connected to a Building Management System via the RS-485 connector. The node  
address switch on each AGM300 may be set from “1” to “15” in order to identify itself to the Building  
Management System. Note that Building Management Systems set to a “0” or “1” address both respond to  
messages from the ADM800 as address “1” therefore you should not have a unit set to “0” and another set to  
“1” on the same network.  
If the AGM300 network is connected directly to a Building Management System it may not be connected to the  
ADM800. However, the ADM800 has two communication ports, an “upstream” port (labeled TO HOST) and a  
“downstream port (labeled TO MONITORS). A BMS node may be connected to the up stream ADM800 port  
while the downstream ADM800 port talks to the AGM300’s. In this case, the BMS is talking “through” the  
ADM800 to the AGM300’s, but not physically on the AGM300/ADM800 network.  
NOTE: If the AGM300 is not at the end of the line in a series connection, then the terminator on the AGM must  
be set to “OUT.” Also, each end of the series must have the terminator set to “IN”.  
MULTIPLE AGM300'S CONNECTED TO A BUILDING MANAGEMENT SYSTEM  
TERMINATOR  
“OUT”  
BMS  
AGM300  
NODE 1  
AGM300  
NODE 2  
AGM300  
NODE 3  
AGM300  
NODE 4  
TERMINATOR  
“IN”  
TERMINATOR  
“OUT”  
TERMINATOR  
“OUT”  
TERMINATOR  
“IN”  
12  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
MULTIPLE AGM300’s CONNECTED TO A BUILDING MANAGEMENT SYSTEM  
DAISY CHAIN FASHION  
BUILDING  
MANAGEMENT  
RS-485  
SYSTEM  
TERMINATOR  
“IN”  
AGM300  
UNIT 1  
AGM300  
UNIT 2  
AGM300  
UNIT 3  
AGM300  
UNIT 4  
NODE 1  
NODE 2  
NODE 3  
NODE 4  
TERMINATOR  
“OUT”  
TERMINATOR  
“OUT”  
TERMINATOR  
“OUT”  
TERMINATOR  
“IN”  
AGM300 - PC Software  
Programming the AGM300 Using PC Software  
NOTE 1: There are two versions of PC software. Use Version 1.0 with AGM300 units with v1.0 Unit Firmware,  
which were built June 2001 or before (as indicated by a serial number that starts with AF01 or lower). Use  
Version 1.52 or higher with v1.21 Unit Firmware or above, which were built August 2001 or later (as indicated  
by a serial number that starts with AH01 or above). The Firmware version is indicated on the label on the ROM  
module inside the AGM300.  
NOTE 2: The PC software uses COM1 by default. Therefore, the interface cable should be connected to the  
port configured as COM1 on the PC. Also, no other software drivers or devices in the PC may control COM1  
when the AGM300 software is in use.  
The connection is made through a standard “straight through” serial port connection. A three-wire connection is  
used (RXD, TXD, and GND). No hardware flow control is used. The AGM300 software automatically  
configures COM1 to match the AGM300 RS-232 communications parameters.  
NOTE 3: Occasionally the laptop connection will not connect properly and only two beeps are heard and the  
program times out. The cure for this is to disconnect the RS-232 cable and cycle power on both the AGM300  
and the laptop. After both the AGM300 and the laptop are up and running connect the RS-232 cable and start  
the software program.  
1. Turn on power to AGM300 and allow it to warm up.  
2. Connect RS-232 interface cable to PC and RS-232 port on the AGM300.  
3. Insert software disk into PC.  
4. Open the AGM300 software using Windows Explorer.  
5. Upon start up, the program will immediately attempt to download data from the AGM300. Several beeps  
can be heard as the program communicates with the AGM300.  
NOTE: To move through the screen use the ARROW keys to move up, down, left and right.  
6. Use the ENTER key to select options and the ESC key to back out of a selection.  
Instruction 3015-4275  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
7. Using the arrow keys go to EDIT; press ENTER – the EDIT Menu DROPS DOWN; select SYSTEM and  
then press ENTER – the AGM LOCATION becomes highlighted. Press ENTER to move to the AGM  
TAG area; use the BACKSPACE key to remove the existing tag; then enter in a new tag. Press  
ENTER to return to LOCATION. Use the ARROW key to select the next item to be addressed. Note  
that you cannot change the “SN” or “FIRMWARE” items. Press the ESC key to return to the menu bar.  
8. Go to EDIT; press ENTER; select ZONES; and then press ENTER. You may now select a specific  
zone to identify and set parameters. When REFRIGERANT is selected use the ARROW keys to move  
up and down through the gas library to locate the gas type for that zone. Highlight the gas type and  
press ENTER.  
9. To set the ALARMS first select EVAC LEVEL; press ENTER; use the BACKSPACE key to clear  
previous setting; and type in the new PPM level. Use the same method to set the SPILL LEVEL and  
LEAK LEVEL.  
10. To close or bypass a zone, set the DISTANCE to 0 feet.  
IMPORTANT: Any time you modify a parameter (zone, system or calibration) and send it to the  
AGM300, please wait for the PC software to indicate that the download is complete before continuing  
with any edits.  
Saving the AGM300 Program to a Disk or Drive  
Go to FILE, PATH, and change the drive letter to the drive where the program is to be saved. Give the program  
a name, such as A:\AGM300 or C:\MY DOCUMENTS\AGM300; press ENTER; go to SAVE SETUP and press  
ENTER again. Type in the file name, such as “CHRM1” (eight characters maximum without a filename  
extension) and then press ENTER. The program will automatically add “.cfg” to the filename that you typed.  
To Open a Saved Program  
Open software; go to FILE, PATH, and clear the old path and type in the proper drive for the saved program  
(A: or C: etc.). Press ENTER; select the proper program; press ENTER again. The saved program will be  
downloaded into the software.  
To Send a Saved Program to an AGM300  
Open the AGM300 PC software; open the proper saved program; connect the PC to the AGM300; and then go  
to AGM, SEND SETUP and press ENTER. The saved program will be sent to the AGM300.  
Trend Data  
Creating a trend data file must be done while connected to the AGM300.  
Go to FILE, PATH, clear the old path and type in the proper drive where the trend data will be saved. Press  
ESC; go to AGM; select GET TREND DATA; select the zone that you want to trend and press ENTER. The  
trend data will appear in a list format. Press ENTER again – a file name screen will appear. Type in the file  
name for that zone (trend1 etc.). Press ENTER and the file is saved as a text file. The text file can be converted  
to a Microsoft Excel file or printed as is.  
NOTE: The TREND file must be saved ZONE by ZONE, with a filename per zone. The file will be saved in  
a notepad format, which can be converted to a delimited Excel file.  
To Convert the TREND Text File to a Microsoft Excel File  
Open Microsoft Excel; open the desired file using All files (*.*); select DELIMITED format and SPACE as the  
delimiter. Select GENERAL as the column data format. The text file now appears as an Excel file. To save the  
file, go to FILE, SAVE AS; select the proper drive; then change the file name extension to “.XLS” and save it.  
Comments or notes may be added to this file as needed and saved.  
To Save and Print the AGM300 Software Screen, Alarm Log,  
Fault Log, or Diagnostic Screen  
Open the software while connected to the AGM300. After the software gets the program, open the desired  
screen. With that screen as the active screen, press the ALT and PRINT SCREEN keys on the PC keyboard.  
Open Microsoft Word; go to FILE, NEW; go to EDIT; then PASTE. The active screen is then brought into Word.  
To save the file, select SAVE AS; select the proper drive and file name; and then press SAVE.  
14  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
USB Type Laptops  
Some of the newer laptops have USB ports and no RS-232 9-pin ports. You will be required to purchase a  
PCMCIA card that provides an RS-232 output. We recommend the PCMCIA card manufactured by:  
SEALEVEL SYSTEMS, INC.  
155 TECHNOLOGY PLACE  
P. O. BOX 830  
LIBERTY, SC 29657  
PHONE – 864-843-4343  
Part number: PC- SIO-232 PCMCIA CARD  
A “ straight through” RS-232 cable and a DB25 to DB9 adapter will be required to connect the laptop to the  
AGM300.  
Instruction 3015-4275  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
AGM300 - Current Loop Interfaces  
Optional 4–20 mAdc Outputs  
Upon installation of the optional 4–20 mAdc Interface Board (Page 7) (P/N 3015-3133), the AGM300 has the  
capability of providing dual 4-20 mAdc scrolling current loop outputs for connection to external monitoring  
devices.  
The interfaces are set up as follows:  
Loop 1 indicates zone area  
Loop 2 indicates PPM  
Connection  
The external devices are connected to the AGM300 using a shielded dual twisted pair cable. Use any of the  
remaining service knockouts to gain access to the interior of the monitor. Locate the Dual 4–20 mAdc Output  
connector (Page 7) and remove it from the circuit board. Secure the wire leads to the connector orienting them  
as shown in the diagram below. Check to make sure the polarity matches the wiring at the external device.  
When you are through securing the connections, carefully plug the connector back onto the circuit board.  
NOTE: When one or both current loop outputs are not used, install a jumper wire between the ground the  
unused loop terminal(s) to prevent the system from generating a current loop fault.  
CAUTION - The loop outputs have isolated grounds.  
Therefore, the cable shield should be terminated at the  
receiver and not at the AGM300. The signal for both  
current loops must be returned to the AGM300.  
Current Loop Connector  
GROUND  
NOTE:  
SIGNAL OUT  
ONLY DO NOT  
APPLY POWER  
LOOP 1  
ZONE  
LOOP 2  
PPM  
NOTE: The ground connections are common.  
LOOP 1 = ZONE  
5 mAdc = Zone 1  
6 mAdc – Zone 2  
---  
LOOP 2 = PPM  
Default  
0.016 mAdc = 1 PPM  
Scalable via Software  
4 mAdc = 0 PPM  
20 mAdc = 1000 PPM  
---  
20 mAdc = Zone 16  
16  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AGM300 - Connecting External Alarms  
Overview  
The AGM300 includes 4 Form C SPDT alarm contacts for the connection of external alarm devices. The relays  
are factory assigned to the following conditions:  
Relay #1  
Relay #2  
Relay #3  
Relay #4  
Leak  
Spill  
Evacuate  
Fault  
Connection  
Use the AC conduit or any of the remaining service knockouts to gain cable access to the interior of the  
monitor. Locate the relay connector (Page 7) and remove it from the circuit board. Secure the leads to the  
connector orienting them as shown in the diagram below. When you are through securing the connections,  
carefully plug the connector back onto the circuit board.  
AGM300 Relay Connector  
NOTE: Each relay may be connected as normally open (NO), or normally closed (NC).  
Latching / Non-Latching – refer to Page 38 to select Alarm Ack Mode.  
Alarm power source – power for the alarm devices connected to the relay contacts may be supplied from the  
Auxiliary AC Power Out Connector, 5 amp. max. (Page 7).  
Instruction 3015-4275  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Typical AGM300 Relay 1 and Relay 2 Wiring  
LEAK  
SPILL EVACUATE FAULT  
Horn  
RS-485  
Aux. Power  
Main AC Input  
Jumper the ‘Neutral’ of the auxiliary power connector to the ‘Common’ on the relay block.  
Connect one end of strobe or horn to the ‘NO’ of whichever level of alarm is appropriate for application.  
The other end of strobe or horn is connected to the ‘Line’ of the auxiliary power connector (5 A max.).  
18  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ADM800 Installation  
ADM800 Oblique Photo  
Upper & Lower Mounting Brackets  
Upper  
Mounting  
Bracket  
Lower  
Mounting  
Bracket  
Instruction 3015-4275  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 - Installation Considerations  
Overview  
Up to four AGM300 monitors may be connected to a single ADM800 display module. The ADM800 can be  
located anywhere up to 4500 feet from the furthest AGM300. The display module should be installed out of  
direct sunlight in a clean area that is not subject to temperature or humidity extremes.  
Warnings and Cautions  
WARNING - Drilling holes in the AGM300 and  
ADM800 enclosure will void the warranty. Please  
use knockouts provided for electrical connections.  
WARNING - Do not operate this equipment in the  
presence of flammable liquids, vapors or aerosols.  
Operation of any electrical instrument in such an  
environment constitutes a safety hazard.  
WARNING - Electrical installation should be  
performed by a certified electrician and should comply  
with all applicable local, state, and federal electrical  
safety codes.  
WARNING - Always de-energize the power supply  
before working on the interior of the instrument.  
CAUTION - The ADM800 contains sensitive  
electronic components that can be easily damaged.  
Be careful not to touch or disturb any of these  
components.  
Inspection  
The ADM800 has been thoroughly inspected and tested prior to shipment from the factory. Nevertheless, it is  
recommended that the instrument be re-checked prior to installation. Inspect the outside of the enclosure to  
make sure there are no obvious signs of shipping damage. Loosen the four screws on the sides of the  
enclosure and remove the front panel. Visually inspect the interior of the display for loose components that may  
have become dislodged during shipment. If damage is discovered, please contact the nearest Bacharach  
Service Center for assistance.  
20  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 - Mounting Instructions  
Screw Locations  
The ADM800 should be installed plumb and level and securely fastened to a rigid mounting surface. The  
enclosure utilizes keyhole mounting brackets designed for #12 pan head fasteners. Locate the four screws as  
shown in the diagram and leave the screw heads protruding approximately 3/16".  
ADM800 Mounting Specs  
Installation  
Hold the display module flat against the mounting surface and allow it to slide down engaging the screw heads  
in the keyhole slots of the mounting brackets. Adjust the screws as necessary to hold the unit securely against  
the mounting surface.  
Instruction 3015-4275  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 - Interior Schematic  
Microcontroller Board  
CPU Reset Switch  
Main Board  
Ground Stud  
AC Power  
Microcontroller  
Board  
CPU Reset  
Switch  
Power  
Supply Board  
120–230  
VAC Input  
Main Board  
AC IN  
Connector T3  
Front Panel  
Chassis  
22  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AC Power  
RS-485 Connectors  
Terminator Switches  
Relay Connector  
Fuses  
1 Amp  
250 VAC  
P/N  
3015-4011  
Terminator  
Switches  
F2  
F1  
TO MONITORS  
TO HOST  
Relay  
Connector  
AC IN  
Connector T3  
RS-485 to  
HGM300  
Monitors  
RS-485 to  
Building  
Management  
System  
Instruction 3015-4275  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 - Electrical Wiring  
It is highly recommended that the ADM800 be connected directly to the AC power source, preferably on its own  
circuit. The connection should be completed with a UL rated multi-conductor wire (14-18 AWG).  
Locate service knockouts on exterior of enclosure and install electrical conduit in the typical manner.  
The ADM800 is available in either a 120 or 230 VAC configuration. The unit’s AC voltage rating is marked on  
the outside of the enclosure. Do one of the following to connect the AC power leads and ground:  
• 120 VAC Installation  
Locate AC IN connector T3 and the Ground Stud on the inside of the enclosure (Page 22).  
Remove the AC IN connector from the circuit board, and then secure the incoming AC power line (black) and  
neutral (white) wires to this connector as shown in the following illustration. Carefully plug the connector back  
onto the circuit board.  
Using the supplied crimp-on ring terminal, washers, and nuts, connect the AC input ground wire (green) along  
with the ground wire connected to the front panel to the unit’s Ground Stud.  
• 230 VAC Installation  
Locate the AC Input Line Filter’s black and white wires, and the Ground Stud on the inside of the chassis  
(Page 22).  
Using wire nuts, secure the incoming AC power Line 1 (black) and Line 2 (white) wires to the AC Input Line  
Filter as shown in the following illustration.  
Using the supplied crimp-on ring terminal, washers, and nuts, connect the AC input ground wire (green) along  
with the ground wire connected to the front panel to the unit’s Ground Stud.  
WARNING - Electrical installation should be performed  
by a certified electrician and should comply with all  
applicable local, state, and federal electrical safety  
codes.  
WARNING - Under no circumstances should this  
instrument be operated without connection to a  
protective ground. Doing so poses a potential shock  
hazard and is also a violation of electrical safety  
standards applicable to this type of equipment.  
A switch or circuit-breaker shall be included in the building installation  
It shall be in close proximity to the equipment and within easy reach of the operator  
It shall be marked as the disconnecting device for the equipment  
24  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 AC Input Power and Ground Connections  
Instruction 3015-4275  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ADM800 - Communication Connections  
AGM300 Network  
The ADM800 Ammonia Display Module is connected to the AGM300 network using a shielded twisted pair  
cable. The maximum distance between the ADM800 and the farthest AGM300 is 4500 feet.  
Use any of the remaining service knockouts to gain access to the interior of the ADM800. The RS-485  
communication wiring between the ADM800 and AGM300 must be connected in the following manner:  
Locate the RS-485 connector marked “TO MONITORS” in the ADM800 (Page 23). This connector is  
located on the far-left bottom of the ADM800 circuit board. Connect one lead of a twisted shielded pair to  
the “B” connection point (the far left point), note the wire color. Connect the second wire to the “A”  
connection point (the middle), note the wire color. Connect the ground to the “GND” connection point.  
Make the wire run to the AGM300 and connect the twisted shielded pair to the AGM300 RS-485 connector  
(Page 7) using the same color code as used on the ADM800.  
Integration with Building Management System  
A second RS-485 connector inside the ADM800 allows a Building Management System to “talk” to the AGM300  
network through the ADM800. The connection is established using a shielded twisted pair cable.  
Use any of the remaining service knockouts to gain access to the interior of the ADM800. Locate the RS-485  
connector marked “TO HOST” (Page 23) and remove it from the circuit board. Secure the wire leads to the  
connector orienting them as shown in the diagram below. Check to make sure that the polarity matches the  
wiring to the Building Management System. When you are through securing the connections, carefully plug the  
connector back onto the circuit board.  
RS-485 Connector  
26  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Changing Terminator Switch Settings  
The terminator switch for the RS-485 “TO MONITORS” connector is shipped from the factory in the terminated  
or “IN” position. This is the correct setting if the ADM800 is connected to only one AGM300, or if it is the last  
device on a network chain. If the ADM800 is being installed in the middle of a network, the terminator must be  
moved to the “OUT” position. Refer to Multiple AGM's on Page 12.  
The terminator switch for the RS-485 “TO HOST” connector is shipped from the factory in the terminated or “IN”  
position. This is the correct setting if the ADM800 is the last device on the network chain. If other devices are  
daisy chained through the ADM800, the terminator must be moved to the “OUT” position. Refer to Connecting  
to a Building Management System on Page 12.  
Locate each terminator switch and determine its position. If it needs to be moved, carefully slide the switch to  
the proper position.  
RS-485  
TO MONITOR  
Connector  
RS-485  
TO HOST  
Connector  
Terminator for RS-485  
Terminator for RS-485  
TO MONITOR  
Connector  
TO HOST  
Connector  
OUT  
OUT  
IN  
IN  
Instruction 3015-4275  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ADM800 - Connecting External Alarms  
Overview  
Two SPDT alarm contacts are available for the connection of external alarm devices. These alarms are useful  
for alerting the user to global conditions anywhere on the network. Each relay can be programmed to respond  
to alarm, fault, or monitor ON conditions (refer to Setting Relay Parameters on Page 36).  
Connection  
Use the AC conduit or any of the remaining service knockouts to gain access to the interior of the ADM800.  
Locate the relay connector (Page 23) and remove it from the circuit board. Secure the wire leads to the  
connector orienting them as shown in the diagram below. When you are through securing the connections,  
carefully plug the connector back onto the circuit board.  
ADM800 Relay Connector  
NOTE: Each relay can be connected as normally open (NO), or normally closed (NC).  
Power for the external alarms can be tapped off the AC IN connector.  
28  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 Operation  
ADM800 Exterior Schematic  
Display Screen  
Navigation Buttons  
Cursor Pad  
LED Indicators  
Display  
Screen  
Navigation  
Buttons  
Cursor  
Pad  
LED  
Indicators  
Instruction 3015-4275  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 - Overview  
The ADM800 displays comprehensive information about the AGM300 network status and enables complete  
programming control of all system operations. The prominent features of the display module are discussed  
below.  
Display Screen  
The display screen is a back-lit LCD.  
Control Keys  
The 4 keys on each side of the display window are used to navigate between screen displays. A graphic box  
that is open to the edge of the window indicates that the adjacent control key is active. The 3 keys along the  
bottom of the window are also used to navigate between screens. A text block directly above the key indicates  
that the key is active. An active key can be used to navigate to another screen. A graphic box, that is not open  
to the edge, displays information only and the adjacent control key is inactive.  
Cursor Key-Pad  
The cursor key-pad is used to move about screens, enter setup information and choose between setup  
parameters. When entering information the LEFT/RIGHT keys move the cursor across an entry field, while the  
UP/DOWN keys scroll data entry options. The ENTER key is used to accept data entries.  
LED Indicators  
The three LED indicators provide an overview of network status.  
GREEN  
RED  
Glows when the system is operational.  
Glows when the ADM800 detects a leak, spill, or evacuate alarm  
condition anywhere on the network.  
YELLOW Glows when the ADM800 detects a system fault anywhere on the network.  
Internal Audible Alarm  
An internal audible alarm is provided that can be programmed to sound when there is an occurrence of the  
selected condition (refer to Audible Alarm on Page 35).  
External Alarms  
Additionally, 2 SPDT alarm contacts are provided for the connection of external alarm devices. These alarms  
are useful for alerting the user to global conditions anywhere on the network. Each relay can be programmed to  
respond to alarm, fault, or monitor ON conditions (refer to Setting Relay Parameters on Page 36).  
30  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 - Screen Displays  
When first powered up a splash screen appears indicating the serial number and firmware revision number.  
After a brief moment the System Screen is displayed. Please refer to the Appendix for a logic diagram of all  
the screen displays.  
There is a 15-minute warm-up period after power is first applied. Under “AGMS”, "WARM UP" is displayed.  
The green indicator on the AGM will blink during the warm-up period.  
IMPORTANT: DO NOT plug the gas-sample hose or inlet port of any zone. The monitor will perform a vacuum  
test during start up and a plugged zone will give a false indication.  
System Screen  
Display  
showing  
“WARM UP”  
Instruction 3015-4275  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 - System Programming  
Setup Parameters  
Before working with the AGM300 network, several global parameters must be defined. Most of these are  
entered in the ADM Setup Screens or via the PC software. Parameters must also be defined for each  
AGM300 monitor connected to the network. These settings serve to identify each monitor and define how they  
operate. These are entered in the AGM Setup Screens. Finally, parameters must be defined for each  
individual monitoring zone. These are entered in the Zone Setup Screens.  
Password Protection  
The ADM800 can be password protected to prevent the unauthorized editing of setup parameters. When  
password protection is enabled, an operator may still navigate between screens to observe settings or monitor  
network status. The password can be alphanumeric.  
The unit is shipped with password protection OFF. Entering a password in ADM Setup Screen #1 enables  
password protection. Once protection is enabled the user will be prompted to enter the password when they  
first attempt to edit a setup parameter.  
To enter a password:  
1. From the top level System Screen press the ADM SETUP key to go to ADM Setup Screen #1.  
2. From ADM Setup Screen #1, press the PASSWORD key.  
3. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to  
modify the individual characters.  
4. Press ENTER to accept the new password or QUIT to revert to the previous setting.  
After entering the password, a 30 minute time period will be allowed for entries; after which, the user will again  
be prompted to re-enter the password.  
IMPORTANT: Please make note and save your password.  
ADM Setup Screen #1  
Password  
32  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 - Setup Programming  
Setting the Clock  
From the top level System Screen press the key adjacent to the date/time field to go to the Clock Setup  
Screen.  
Clock Setup Screen  
Use the LEFT/RIGHT cursor keys to move the cursor across the field to the value you wish to edit. Use the  
UP/DOWN cursor keys to modify the field value and press the ENTER key to accept your entries. When you are  
through editing the date/time value, do one of the following:  
Press the SET AGM CLOCKS key to synchronize all AGM300 monitors to the current date/time setting.  
Press the QUIT key to reject all edits and revert to the previous setting.  
Press the SET ADM CLOCK key to accept the current date/time setting.  
Instruction 3015-4275  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Navigating to the 1st ADM Setup Screen  
From the top level System Screen press the ADM SETUP key to go to ADM Setup Screen #1.  
ADM Setup Screen #1  
Contrast  
This value sets the screen brightness.  
1. Press the CONTRAST key to adjust its setting.  
2. Use the UP/DOWN cursor keys to modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Number of AGM Units  
This is the number of AGM300 monitors from 1–4 connected to the ADM800.  
1. Press the #OF AGMS ON NET key to adjust the number.  
2. Use the UP/DOWN cursor keys to modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
NOTE: When first powered up, the ADM800 will automatically detect all AGM300s on the network with  
unique node addresses. Refer to the heading AGM Node Address Dip Switch Settings on Page 11.  
AGM Baud Rate  
This is the baud rate of all AGM300 monitors connected to the network.  
1. Press the AGM BAUD key to adjust the baud rate.  
2. Use the UP/DOWN cursor keys to scroll the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
4. The default valve is 19200.  
NOTE: All AGM300s on the network must have the same baud rate.  
34  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Audible Alarm  
This parameter selects the function of the ADM800’s internal audible alarm.  
1. Press the AUDIBLE ALARM key to adjust the setting.  
2. Use the UP/DOWN cursor keys to select unused, monitor on, evacuate, spill, leak, fault, or alarm.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Enabling Building Management System Connection  
This enables or disables the connection to a Building Management System.  
1. Press the BMS key to adjust the setting.  
2. Use the UP/DOWN cursor keys to toggle the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Building Management System Stop Bits  
This value is the number of stop bits for the ADM800 on a Building Management System.  
1. Press the BMS STOPBITS key to adjust the setting.  
2. Use the UP/DOWN cursor keys to modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Building Management System Baud Rate  
This is the baud rate of the Building Management System.  
1. Press the BMS BAUD key to adjust the setting.  
2. Use the UP/DOWN cursor keys to modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Password  
This field is used to define a system password. The default setting is 000, which provides no password  
protection.  
1. Press the PASSWORD key to enter a password.  
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to  
modify the individual characters.  
3. Press ENTER to accept the new password or QUIT to revert to the previous setting.  
Instruction 3015-4275  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Navigating to the 2nd ADM Setup Screen  
From ADM Setup Screen #1 press the MORE key to go to ADM Setup Screen #2. Press the BACK key to  
return to ADM Setup Screen #1.  
ADM Setup Screen #2  
Setting Relay Parameters  
This setting determines the function of relays 1 and 2 mounted inside the ADM unit.  
1. Press the key adjacent to the relay parameter you want to adjust.  
2. Use the UP/DOWN cursor keys to select unused, monitor on, evacuate, spill, leak, fault, or alarm.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
AGM Node Address  
This value is the node address from 1 to 15 assigned to each AGM300 that is on the network. These values  
must match the DIP switch settings on each AGM300 circuit board. Refer to the heading AGM Node Address  
Dip Switch Settings on Page 11.  
1. Press the key adjacent to the AGM# NODE you want to set.  
2. Use the UP/DOWN cursor keys to modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Node Address DIP  
Switches in AGM Unit  
36  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AGM300 - Setup Programming  
Navigating to the 1st and then 2nd AGM Setup Screen  
From the System Screen, press the key adjacent to the target AGM300 to go to that unit's AGM Setup  
Screen #1.  
AGM Setup Screen #1  
From AGM Setup Screen #1 press the SETUP key to go to AGM Setup Screen #2. Press the QUIT key to  
return to the System Screen.  
AGM Setup Screen #2  
Instruction 3015-4275  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Location  
This is the name you assign to the AGM300 to identify its location. It may have up to 12 alphanumeric  
characters.  
1. Press the LOCATION key to modify the setting.  
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to  
modify the individual characters.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Number of Zones Installed  
IMPORTANT: Do not change the number of zones to deactivate excess zones. Instead, go to the individual  
zone that you wish to disable and set its distance to zero (refer to Navigating to the 1st Zone Setup Screen on  
Page 41). The monitor will perform a self-check on power-up, and if the number of zones detected does not  
agree with the number of zones installed a fault will occur.  
1. Press the NUM ZONES INSTALLED key to adjust the setting.  
2. Use the UP/DOWN cursor keys to modify the number.  
3. Press ENTER to accept the new number or QUIT to revert to the previous setting.  
NOTE: The AGM300 auto setting detects the number of zones present on power-up. DO NOT PLUG ANY  
ZONES. Plugging a zone will give the monitor a false indication during power up. The number detected on  
power up is the default value for this parameter.  
Alarm Ack Mode  
This function programs the relays in the AGM unit for latching or non-latching operation.  
1. Press the ALARM ACK MODE key to adjust the setting.  
2. Use the UP/DOWN cursor keys to toggle between settings.  
AUTO  
– Non-latching (Alarm relay will automatically de-energize when the gas level drops below  
its alarm point.)  
MANUAL – Latching (Alarm relay remains energized, and will not release until the alarm condition has  
been manually acknowledged. Refer to Acknowledging Alarms on Page 52.)  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Audible Alarm  
This parameter selects the function of the AGM300’s internal audible alarm device.  
1. Press the AUDIBLE ALARM key to adjust the setting.  
2. Use the UP/DOWN cursor keys to select unused, monitor on, evacuate, spill, leak, fault, or alarm.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Zone Hold Time  
Sets the length of time a zone will be monitored when the zone hold feature is actuated (see Zone Hold Mode  
on Page 45). The default is 15 minutes. This parameter can be set to any value between 1 and 999 minutes.  
1. Press the ZONE HOLD MINUTES key to adjust this setting.  
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to  
modify the individual numbers.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Detection Limit  
This is essentially a squelch setting that instructs the instrument to interpret PPM readings below the  
designated level as 0. You can set this value from 1 to 99 PPM.  
1. Press the DETECTION LIMIT key to adjust this setting.  
2. Use the UP/DOWN cursor keys to modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
38  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Loop 2 Factor  
This sets the PPM scale factor for current loop number 2. To calculate the current output, multiply the scale factor  
by the PPM and add 4. For example, at the default scale factor of 0.016, a measurement of 100 PPM would  
generate a current of 5.6 mA. In no case can current output exceed the 20 mA capacity of the interface.  
1. Press the LOOP2 FACTOR key to adjust the setting.  
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to  
modify the individual numbers.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Re-Zero Mode  
This parameter defines the frequency at which the instrument re-zeros the optical sensor.  
1. Press the REZERO MODE key to adjust the setting.  
2. Use the UP/DOWN cursor keys to toggle between settings.  
AUTO  
– Sets the instrument to re-zero every 10 minutes.  
ZONE CHANGE – Sets the instrument to re-zero at each zone change. This is the most accurate  
setting but increases the time interval between measurement cycles.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Instruction 3015-4275  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Navigating to the 3rd AGM Setup Screen  
From AGM Setup Screen #2, press the MORE key to go to AGM Setup Screen #3. Press the BACK key to  
return to AGM Setup Screen #2.  
AGM Setup #3 Screen  
Service Timeout  
Sets the length of time the Service Mode is in effect. Stops the monitor for up to 240 minutes (4 hrs) to allow for  
servicing the unit. See Service Mode on Page 61.  
1. Press the SERV TIMEOUT key to adjust the setting.  
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to  
modify the individual numbers.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
40  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 - Zone Setup Programming  
Navigating to the 1st Zone Setup Screen  
There are two ways to display Zone Setup Screen #1:  
• From the System Screen (Page 31) press the key adjacent to the AGM300 unit associated with the zone you  
wish to work with. Next, press the ZONES key; and then press the key adjacent to the desired zone.  
• From the top level Zone Screen (Page 44) press the key adjacent to the zone you wish to work with. To  
select a different zone, use the UP/DOWN cursor keys or the NEXT key to scroll the screen display.  
Zone Setup Screen #1  
At Zone Setup Screen #1, use the LEFT/RIGHT cursor keys to switch the display to a different zone.  
Location  
This is the name you assign to the monitoring zone. It may have up to 12 alphanumeric characters.  
1. Press the LOCATION key to enter or modify the name of the zone.  
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to  
modify the individual characters.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Distance  
This is the tubing length in feet from the AGM300 to the termination of the gas sample intake line. This value  
may range from 0 to 500 feet. The default value is 100 feet. Note that setting this value to “0” causes the zone  
to be bypassed during normal operation, and the word “DISABLED” in appear in place of the zone’s PPM  
reading in the Zone Screen (Page 44).  
1. Press the DISTANCE key to adjust the distance value.  
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to  
modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Instruction 3015-4275  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Average Temperature  
This temperature reading may be changed to the exact temperature at the sample point for greater accuracy.  
The adjustment range is from –50 to +70 degrees C.  
1. Press the AVG TEMP key to adjust the temperature value.  
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to  
modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Current PPM  
This box displays the last PPM measurement recorded in this zone. You cannot modify this value.  
Log Interval  
Sets the length of time between entries in the trend log. The default is one entry every 1440 minutes (24 hours).  
A value of 000 indicates that each measurement for a given zone is logged into the trend log.  
1. Press the LOG INTERVAL key to adjust the log interval value.  
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to  
modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Navigating To the 2nd Zone Setup Screen  
From Zone Setup Screen #1 press the MORE key to go to Zone Setup Screen #2. Press the BACK key to  
return to Zone Setup Screen #1.  
Zone Setup Screen #2  
Leak Level  
This is the concentration level in PPM that will activate a leak alarm condition.  
1. Press the LEAK LEVEL key to adjust the value.  
2. Use the UP/DOWN cursor keys to modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
NOTE: This value must be less than the spill level.  
42  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Spill Level  
This is the concentration level in PPM that will activate a spill alarm condition.  
1. Press the SPILL LEVEL key to adjust the value.  
2. Use the UP/DOWN cursor keys to modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
NOTE: This value must be less than the evacuate level and greater than the leak level.  
Evacuation Level  
This is the concentration level in PPM that will activate an evacuate alarm condition.  
1. Press the EVAC LEVEL key to adjust the value.  
2. Use the UP/DOWN cursor keys to modify the setting.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
NOTE: This value must be greater than the spill level.  
Re-Setting the Peak PPM Value  
Pressing this key resets the current peak PPM level stored in memory and displays it at the top of the screen.  
Instruction 3015-4275  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General Operation  
Functional Overview  
Normally each AGM300 will sequentially perform measurements on its active zones without user input. The  
total time it takes an AGM300 to complete a measurement cycle is directly proportional to the number of active  
zones and the physical length of the sample-gas lines. Monitors linked together on a network operate  
independently of each other and consequently complete their respective measurement cycles at different rates.  
The ADM800 operates by polling the network approximately once every 12 seconds to determine the current  
status of each AGM300 monitor. If more than one AGM300 is connected to the network, it will sequentially poll  
each monitor for its status. As a practical matter this simply means that the more complicated the network, the  
longer it will take the ADM800 to update the status information for all zones.  
To display all the zones being monitored, go to the System Screen (page 31) and press the ZONES key.  
Working with the Zone Screen  
The top level Zone Screen provides a summary view of zone activity across the network. Each screen displays  
information from a single AGM300 monitor as indicated at the top of the display. Each screen can display  
information for up to 8 zones, and depending on the configuration of the system, there can be up to two screens  
for each AGM300. Use the UP/DOWN cursor keys or the NEXT key to scroll the screen display.  
Each box displays the zone location and the last PPM measurement. A blinking box indicates that a  
measurement is currently being taken in that location. An inverse flashing box indicates an alarm condition in  
that zone.  
Use the LEFT/RIGHT cursor keys to scroll through the zones.  
Zone Screen  
To further investigate the status of any zone, press the adjacent key to go to the Zone Setup #1 Screen  
(Page 41). You may also press the SYSTEM key to go to the System Screen (Page 31).  
44  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Zone Hold Mode  
A zone can be placed on hold and continuously monitored for a length of time determined by the Zone Hold  
Time value (Page 38). To place a zone on hold, press and hold down the key adjacent to that zone until you  
hear a beep. The word "HOLDING" appears to identify which zone is being held.  
Zone Screen (Hold Mode)  
Zone on Hold  
Once in the hold mode you can further investigate the status of the zone by pressing the adjacent key and going  
adjacent key until the unit beeps and the screen display reverts to normal. You may also press and hold down  
the key adjacent to any other zone to release the current zone and put the new zone on hold.  
Alarm Conditions  
When an alarm condition is detected anywhere on the network the red ALARM LED will glow. Additionally, an  
external alarm device may activate and an audible alarm may sound if those features have been enabled  
(Pages 28 & 35).  
An inverse flashing box indicates an alarm condition in the affected zone.  
Zone Screen (Alarm Mode)  
Zone in Alarm  
Instruction 3015-4275  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
If the current Zone Screen appears normal, scroll through the displays using the NEXT/BACK keys or the  
UP/DOWN cursor keys until you locate the affected zone. You can further investigate the status of the affected  
zone by pressing the adjacent key and going to the Zone Setup #1 Screen (Page 41). You may also press the  
ALARM key to go directly to the Alarm Summary Screen (Page 50). See Working with Alarms on Page 50 to  
learn more about responding to alarms.  
Fault Conditions  
If a system malfunction occurs, the yellow FAULT LED will glow. Additionally, an external alarm device may  
activate and an audible alarm may sound if those features have been enabled (Pages 28 & 35).  
From the top level Zone Screen (Page 44), press the SYSTEM key to go to the System Screen (Page 31) and  
then press the FAULT key to go to the Fault Screen (Page 54). See Working with System Faults on Page 54 to  
learn more about responding to system faults.  
46  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Working with the System Screen  
The System Screen provides a summary view of the entire AGM300 network. The boxes on the left side of the  
screen indicate the status of each AGM300 monitor including the name, the current zone, and whether fault or  
alarm conditions are present on the network.  
System Screen  
Alarm Conditions  
When an alarm condition is detected anywhere on the network the red ALARM LED will glow. Additionally, an  
external alarm device may activate and an audible alarm may sound if those features have been enabled  
(Pages 28 & 35).  
The ALARM box in the upper right hand corner of the screen will blink and the count value will change to  
indicate the total number of alarm conditions on the network. To investigate an alarm, press the ALARMS key to  
go to the Alarm Summary Screen (Page 50). See Working with Alarms on Page 50 to learn more about  
responding to alarms.  
System Screen (Alarm Mode)  
Alarm  
Condition  
Instruction 3015-4275  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Alarm Log  
From the System Screen press the key adjacent to the AGM300 unit that you wish to view.  
AGM Setup #1 Screen  
Press the ALARM EVENT LOG key to display Alarm Log Screen.  
Alarm Log Screen  
Use the LEFT/RIGHT and UP/DOWN cursor keys to move through the log.  
The Alarm log shows the last 20 alarm events. An alarm event is the occurrence of any alarm, any change in  
alarm level, or the clearing of any alarm. The alarm level is indicated by an L (leak), S (spill), or E (evacuate).  
The time and date of the alarm are displayed when the cursor is placed over the alarm.  
48  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fault Conditions  
If a system malfunction occurs, the yellow FAULT LED will glow. Additionally, an external alarm device may  
activate and an audible alarm may sound if those features have been enabled (Pages 28 & 35).  
The FAULT box on the right of the screen will blink and then counts the number of AGM units on the system  
with a fault. To investigate a fault, press the FAULTS key to go to the Fault Screen (Page 54). See Working  
with System Faults on Page 54 to learn more about responding to system faults.  
System Screen (Fault Mode)  
Fault  
Condition  
Instruction 3015-4275  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Working with Alarms  
Functional Overview  
If the PPM level for any zone exceed its designated spill, leak, or evacuate thresholds, an alarm condition will  
be created. Once the AGM300 completes a measurement cycle in the affected zone the alarm condition will be  
indicated. At that time the red ALARM LED on the AGM300 will glow. Additionally, an external alarm device  
may activate and an audible alarm may sound if those features have been enabled (Pages 17 & 38).  
The next time the ADM800 polls the affected monitor its red ALARM LED will glow. Additionally, an  
external alarm device may activate and an audible alarm may sound if those features have been enabled  
(Pages 28 & 35).  
The alarm condition may or may not have to be acknowledged for the system to return to normal operation.  
Responding To Alarms  
There are two screens from which an operator can respond to alarms:  
• From the System Screen (Page 31) press the key adjacent to the blinking ALARMS field to go to the Alarm  
Summary Screen.  
• From the top level Zone Screen (Page 44) press the ALARMS key at the bottom of the display window to go  
to the Alarm Summary Screen.  
Alarm Summary Screen  
The Alarm Summary Screen displays a list of all alarm conditions pending across the network. The screen is  
divided into 8 boxes, with each box representing a single alarm. If more than 8 alarms are pending additional  
pages can be displayed by pressing the MORE key.  
Each box displays the zone number, zone name, and the current PPM reading. A flashing box indicates an  
alarm that has not been acknowledged. A static box represents an alarm that has been acknowledged but has  
not yet been cleared from the system.  
50  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Alarm Detail Screen  
To further investigate an alarm, press the adjacent key to go to the Alarm Detail Screen.  
Alarm Detail Screen  
The Alarm Detail Screen displays more comprehensive information about the nature of the alarm including:  
• Complete location information  
• Current PPM  
• Peak PPM and peak time  
• Type of alarm, alarm time, and date  
This screen includes three active navigation keys at the bottom of the display.  
ACK  
Acknowledges the alarm as described in the next section  
QUIT  
Returns you to the Alarm Summary Screen  
SETUP Takes you to the Zone Setup #1 Screen (Page 41), which enables you to review the zone setup  
parameters and access the Trend Screen.  
Instruction 3015-4275  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Acknowledging Alarms  
Each pending alarm may require, depending upon selected alarm mode, acknowledgment before the system  
returns to normal operation (refer to Alarm Ack Mode on Page 38). To acknowledge an alarm, navigate to the  
Alarm Detail Screen and press the ACK key as previously described. You will then be returned to the Alarm  
Summary Screen and the box associated with that alarm will no longer be blinking, indicating that the alarm  
has been acknowledged. Repeat this procedure to acknowledge any remaining alarms.  
Alarm Summary Screen (Acknowledge Mode)  
Once all the alarms associated with a given AGM300 are acknowledged, its red ALARM LED will turn off and  
any external alarms connected to the AGM300 relays will de-activate. All pending alarms across the entire  
network must be acknowledged before the ADM800 returns to normal operation. Once that occurs, its red  
ALARM LED will turn off and any associated external alarms connected to the ADM800 relays will de-activate.  
Keep in mind that the system will continue to generate new alarms if PPM values above the alarm thresholds  
are detected.  
52  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Working with the Trend Screen  
Overview - Log Interval  
The AGM300 retains a data log of 100 measurements for each zone. The log interval is the number of minutes  
from 1 to 1440 between each log point. This parameter can be changed from the Zone Setup #1 Screen  
(Page 41).  
The default setting for this parameter is 1440 minutes (24 hours). If the log interval time is set to 0, then a  
measurement is recorded in the trend log after every measurement cycle. Therefore, after the trend log is filled it  
will contain the last 100 measurement points for a zone. If you want the data logged less frequently you should  
increase this value. It is important to remember that cycle time is dependent on many factors, including the  
number of zones being monitored, input line length, and the run zeroing mode selected. Before changing this  
value it may be useful to first review the log data using the Trend Screen to determine the nominal cycle time.  
Navigating To the Trend Screen  
Access the Trend Screen from either the System Screen or Zone Screen as follows:  
• From the System Screen (Page 31) press the key adjacent to the AGM300 unit associated with the zone you  
wish to work with to display that monitor's AGM Setup Screen #1. Next, press the ZONES key to go to Zone  
Setup Screen #1, and then press the TREND key at the bottom of the display to go to the Trend Screen.  
• From the top level Zone Screen (Page 44) press the key adjacent to the zone you wish to work with to first  
display its Zone Setup Screen #1. Then press the TREND key at the bottom of the display to go to the Trend  
Screen.  
Trend Screen  
The trend graph opens with the cursor located over the most recent data point. Use the LEFT/RIGHT cursor  
keys to move the cursor to different data points. Holding a key down will cause cursor to move rapidly across  
the screen. As you move the cursor position, the date and time of that reading, along with the PPM value, are  
displayed at the top of the screen above the graph. The trend graph is automatically scaled to accommodate  
the largest PPM value displayed. The ZOOM OUT and ZOOM IN keys allow you to adjust the vertical scale of  
the graph.  
The trend data is stored on a first-in, first-out basis. After 100 trend values have been stored the 101st value will  
replace the first value stored. Therefore, in normal operation, when entering the trend mode the cursor will be  
located at the most recent data point. The data points to the left of the initial cursor location will be the next  
most recent. The data point to the right will be the oldest data point in the buffer and will be over written by the  
next data point.  
Instruction 3015-4275  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Working with System Faults  
Functional Overview  
If a system malfunction occurs, the AGM300 will detect the problem and cause its yellow FAULT LED to glow.  
Additionally, an external alarm device may activate and an audible alarm may sound if those features have  
been enabled (Pages 17 & 38).  
The next time the ADM800 polls the affected monitor its yellow FAULT LED will also glow. Additionally, an  
external alarm device may activate and an audible alarm may sound if those features have been enabled  
(Pages 28 & 35).  
Depending on the nature of the fault, the AGM300 may or may not continue to operate normally. Under a non-  
critical fault condition, the AGM300 will continue to measure and log data, but some peripheral functions may  
be compromised. Under a critical fault condition, action is required for the AGM300 to operate normally. The  
table on the following page lists the various fault conditions and explains what action should be taken to correct  
the problem.  
Navigating to the Fault Screen  
First, display the System Screen (Page 31). Next, press the key adjacent to the target AGM300 unit to go to  
that monitor’s AGM Setup Screen #1 (Page 37). Finally, press the FAULTS key to display the Fault Screen of  
the selected monitor.  
Fault Screen  
54  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
CRITICAL FAULTS  
NO FLOW ON ZONE – Go to the System Screen (Page 47) and press the ZONES key. The Zone  
Screen (Page 44) will display a “NO FLOW” message in each individual zone affected. Check for a  
blockage in the air sample line or at the line end filter. Once the blockage has been cleared, the  
AGM300 will return to normal operation after the zone has been sampled. NOTE: This can take some  
several minutes since it is dependent upon how many zones there are and their lengths. The AGM300  
will clear the fault the next time it polls the effected zone and the ADM800 will return to normal  
operation the next time that it polls the AGM300.  
NO FLOW ON PURGE – Check the purge line for a blockage. Verify that the length of the purge line  
and exhaust line do not exceed 300 feet in length.  
NO FLOW DETECTED – Check for proper pump operation. Check the water trap – drain if necessary.  
CLIPPING FAULT – The detector voltage may be out of tolerance. Check the Diagnostic Screen  
(Page 59) for the DET voltage, AVE voltage and ZERO voltage. Call the factory with this information for  
further instructions.  
REZERO VOLT TOL – The detector output voltage is out of tolerance. Check the Diagnostic Screen  
as in item 4 and contact the factory for assistance.  
TRIGGER FAULT – No trigger from IR source pulser. Contact factory with all information from the  
Diagnostic Screen for further instructions.  
NON CRITICAL FAULTS  
BOX TEMP FAULT – Enclosure’s internal temperature is outside normal range (or IR sensor has  
failed). Check the installation to verify that the monitor is not being subjected to extreme temperatures.  
Verify that the ventilation holes are not obstructed. Check the Diagnostic Screen for the ZERO  
temperature, BNCH temperature and BOX temperature. Call the factory with this information for further  
instructions.  
BENCH TEMP FAULT – Optical bench is outside of normal operating range (or sensor has failed).  
Check the installation to verify that the monitor is not being subjected to extreme temperatures. Check  
the Diagnostic Screen for the ZERO temperature, BNCH temperature and BOX temperature. Call the  
factory with this information for further instructions.  
PRESSURE SENSOR – Manifold pressure is outside normal operating range (or sensor has failed).  
Check the Diagnostic Screen and record ALL data. Call the factory with this information for further  
instructions.  
LOOP FAULT – This would only be displayed if the dual 4–20 mA option was installed and one or both  
current loops are open. Check the wiring to load/monitoring circuit on both 4–20 mA loops.  
CONFIG FAULT – There is an error in AGM Setup Screen #2 – Number Zones Installed field, or in  
ADM Setup Screen #1 – Number of AGMS on Network field. Check that the number of zones installed  
for each AGM unit and the number of AGM units on the network are properly programmed. Check to  
ensure that the manifold solenoid cable connector in each AGM unit is securely fastened to its terminal  
connector. Check for an illegal parameter. If necessary, reset to the factory default settings.  
Reset to Factory Default Settings  
IMPORTANT: Performing this function wipes out all program parameters, alarms, faults, trends and  
log files.  
Resetting the AGM300 – Press and hold down the Factory Default switch inside the AGM300 (Page 7);  
cycle the AC power OFF and then ON; listen for five beeps; and then release the switch. Reprogram the  
AGM300 per Section AGM300 – Setup Programming on Page 37.  
Resetting the ADM800 – Occasionally it will be necessary to rest the ADM800 to factory default settings.  
From the System Screen (Page 31), press and hold the key adjacent to the ALARMS function; cycle AC  
power OFF and then ON; listen for five beeps; and then release the key.  
Instruction 3015-4275  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Clearing System Faults  
If the fault condition is associated with an AGM300, the monitor will return to normal operation soon after the  
problem is corrected. At that time the YELLOW LED will turn off and any external alarms connected to the  
monitor’s alarm relays will also turn off. The ADM800 will return to normal operation the next time it polls the  
affected AGM300 monitor.  
Once the system malfunction has been corrected it may take some time for the fault condition to clear  
completely. If the fault is associated with a specific zone, the AGM300 must first cycle back to the affected  
zone before it returns to its normal operating mode. At that time the yellow FAULT LED will turn off and any  
external alarms connected to the AGM300 unit will de-activate. The ADM800 will return to normal operation the  
next time it polls the affected monitor.  
Viewing Fault Log  
A data log of the last 20 fault conditions is retained in memory. From the Fault Screen (Page 54) press the  
LOG key to view a display of the fault log.  
Fault Log Screen  
This screen lists potential fault conditions in the left column and displays a check mark indicating which  
problems were associated with each fault condition as represented by the vertical cursor bar. As you move the  
bar horizontally using the LEFT/RIGHT cursor keys, the date and time the fault condition was detected is  
displayed in the upper right hand corner of the display window.  
Anytime the fault status changes, there is an entry in the fault log, both when the fault occurs and when it is  
cleared.  
56  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Working with the Calibration Screen  
Overview  
The Calibration Screen is used to adjust the IR sensor’s calibration factor.  
IMPORTANT: Changing information on CAL FACTORS will VOID the factory calibration. Typically, the unit will  
remain within the factory-calibrated accuracy indefinitely and no calibration is required. Complex software  
algorithms adjust for temperature drift, IR source aging, and pressure changes in order to keep the unit within  
factory accuracy specifications.  
Navigating to the Calibration Screen  
First, display the System Screen (Page 31). Next, press the key adjacent to the target AGM300 unit to go to  
that monitor’s AGM Setup Screen #1 (Page 37). Finally, press the CAL key to go to the selected monitor’s  
Calibration Screen.  
Calibration Screen  
Adjusting Calibration Factor  
The factory default cal factor for standard units is 1.000.  
Proceed as follows to adjust the current calibration factor:  
1. Press the CAL FACTOR key to edit the value.  
2. Use the LEFT/RIGHT cursor keys to move across the entry field and the UP/DOWN cursor keys to  
modify the individual numbers.  
3. Press ENTER to accept the new entry or QUIT to revert to the previous setting.  
Instruction 3015-4275  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Procedure  
The CAL FACTOR is determined by sampling a known dilution of ammonia gas. The sample must be prepared  
to less than half the desired accuracy, and the concentration must be corrected for ambient temperature and  
pressure at the time of measurement.  
Calibration is best performed at or near full scale (10,000 PPM). It can, however, be done at any concentration,  
and ideally in the range where maximum accuracy is desired down to, but not below, 100 PPM.  
A cylinder of ammonia gas at a certified PPM level must be used to assure sampling occurs at ambient  
conditions. A minimum sample size of 5 liters is required.  
The AGM should be operating for at least one hour prior to performing a calibration.  
Prepare the AGM300 for sampling by initially setting its CAL FACTOR to 1.000 (Page 57). Next, set up the  
AGM300 for a logging interval of zero minutes (Page 42), and place the AGM300 in its zone hold mode for the  
zone you wish to set up (Page 45).  
Connect the sample bag directly to the intake port for the zone you have set up and allow the AGM300 to  
sample the entire bag. When sampling is complete view the trend data for the zone used to sample (Page 53).  
Read the measured PPM by placing the cursor on the spikes cause by the sample. If the bag was large enough  
for multiple samples, average the most stable ones.  
The new CAL factor is computed by dividing the known gas value by the measured value. Typically this value  
will be between 0.95 and 1.05. Enter the computed CAL factor into the monitor using the ADM800 (Page 57) or  
PC interface. This value is stored in non-volatile memory.  
58  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Working with the Diagnostic Screen  
Overview  
The Diagnostic Screen displays reference values for use by repair technicians for troubleshooting purposes.  
Navigating to the Diagnostic Screen  
Display the System Screen (Page 31); press the key adjacent to the target AGM300 unit to go to that  
monitor’s AGM Setup Screen #1 (Page 37); and then press the DIAG key to go to the selected monitor’s  
Diagnostic Screen.  
Diagnostic Screen  
Diagnostic Screen Overview  
The Diagnostic Screen contains sensor data and status information useful for troubleshooting various fault  
conditions. An explanation of each line is given below along with normal operating ranges.  
LINE 1: Detector Voltage – This is the peak-to-peak output of the IR sensor, in the absence of ammonia  
gas this value can range from 3.900V to 4.500V.  
LINE 2: Average Detector Voltage – This is simply a running average of the values displayed in line 1.  
LINE 3: Zero Voltage – This is the IR sensor output that was stored during the last purge cycle and has  
the same range as line 1.  
LINE 4: Noise – This is a 16 point running average of the noise portion of IR bench output. This reading is  
valuable mainly when ammonia gas is NOT present.  
LINE 5: Average Absorbency – This is the optical absorbency computed from the values in lines 2 and 3.  
In the absence of ammonia gas the absorbency is 0.000AU. When sampling gas, its value varies  
proportionally with the gas concentration.  
LINE 6: uMoles/L – This is the absolute concentration in micro-moles per liter of ammonia gas based on  
line 4 and the internal calibration. There are two figures given. The first (which is annotated by a B) is the  
actual measurement at the IR bench. The second is the calculated value corrected to ambient conditions  
(temperature + pressure).  
Instruction 3015-4275  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
LINE 7: PPM – Parts Per Million is the volume concentration referenced to standard temperature and  
pressure and is computed from lines 5, 8 and 13. There are two figures given. The first (annotated by a B)  
is the actual PPM at the IR bench. The second is a PPM reading normalized to standard temperature and  
pressure.  
LINE 8: Zero Temperature – This line contains the sensor and enclosure temperature measured and  
stored during the last purge cycle in degrees C.  
LINE 9: Bench Temperature – This is the current IR sensor temperature in degrees C as well as the raw  
voltage coming from the temperature sensor itself. This value can range from ambient to ambient +15  
degrees C.  
LINE 10: Box Temperature – This is the current internal enclosure temperature along with the raw  
voltage from the temperature sensor, and has the same range as line 9.  
LINE 11: Pressure – This is the current absolute manifold pressure in PSIA along with the output voltage  
of the pressure sensor. This value should always be 0.2 to 1.0 PSIA below ambient (line 13).  
LINE 12: Vacuum – The vacuum pressure is measured every purge cycle by closing all sample valves.  
Its value is typically 2.5 to 4.0 PSIA below ambient (line 13).  
LINE 13: Ambient – Ambient pressure is measured every purge cycle with the sample pump off and the  
manifold open. Its value is weather and altitude dependent and can range from 10.0 to 15.5 PSIA.  
60  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Mode  
When activated, the Service Mode will disable a specific AGM300 unit for a specified length of time. The default  
is 60 minutes. This time interval can be changed per Section Service Timeout on Page 40.  
IMPORTANT: Note that while in the Service Mode no measurements are made, nor are any alarms activated.  
From the System Screen (Page 31), select the key adjacent to the AGM unit you wish to place into its Service  
Mode. From AGM Setup Screen #1 (Page 37) press the SERVICE MODE ENTRY key twice within  
3 seconds.  
Note that the System Screen will show what monitors are in the Service Mode.  
To exit the Service Mode, again press the SERVICE MODE ENTRY key twice within 3 seconds.  
AGM Setup Screen #1  
Press this key  
twice within  
3 seconds to  
place the  
selected AGM  
unit into its  
Service Mode  
System Screen (Service Mode)  
Monitor in  
Service  
Mode  
Instruction 3015-4275  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Notes:  
62  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Instruction 3015-4275  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AGM300 / ADM800 Maintenance  
Warnings and Cautions  
WARNING - Always remove AC input power to the monitor  
before working on the interior of the instrument.  
CAUTION - When working inside the instrument be very  
careful not to dislodge any electrical wiring or pneumatic  
tubing. The AGM300 contains sensitive electronic  
components that can be easily damaged. Be careful not to  
touch or disturb any of these components.  
Servicing Inline Filter  
Locate the inline filter (Page 6) and inspect it for obvious signs of contamination. If the filter is contaminated or  
has started to turn red in color it should be replaced with P/N 3015-3071.  
To replace the filter, remove it from the retaining clip and carefully slide the two pieces of tubing off the fittings.  
Re-fit the tubing on the new filter, making sure the flow arrow points to the right. Complete the installation by  
pushing the new filter back into the retaining clip.  
Servicing Air Lines  
System air lines should be checked periodically for obvious signs of kinks, damage, and contamination.  
Replace the tubing as required with P/N 304-2742 or equivalent.  
The end-line filters should be checked periodically and replaced when there are obvious signs of contamination.  
To service the filter simply remove it from the air line and replace it with P/N 3015-3420.  
Replacing Fuses  
Locate the fuse holders (Pages 7 & 23) and remove the fuses from their holders. Visually inspect and test the  
fuses with an ohm meter for continuity. If necessary, replace any defective fuses.  
AGM300, 120 VAC Input, Power Supply P/N 3015-3167  
F1: 0.5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2538  
F2: 250 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2537  
F3: 5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2539  
F4: 63 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2535  
AGM300, 230 VAC Input, Power Supply P/N 3015-3161  
F1: 250 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2537  
F2: 125 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2536  
F3: 5 A, 250V Fast Acting, 5mm x 20mm, P/N 604-2539  
F4: 32 mA, 250V Fast Acting, 5mm x 20mm, P/N 604-2534  
ADM800  
F1 & F2: 1 A, 250VAC, Fast Acting, 5mm x 20mm, P/N 3015-4011  
64  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Optional Accessories  
Intake Manifolds  
Up to three additional intake manifolds (P/N 3015-3419 & P/N 3015-3411) can be installed in the AGM300 to  
increase the total capacity to 16 zones. Each manifold requires a sample line-end filter for each of the 4 zones.  
Please follow the instructions provided to install the manifolds. Then follow the instructions earlier in this manual  
to install the air lines and complete the zone programming steps. The AGM300 setup must also be updated to  
recognize the new zones.  
4–20 mA Interface Board  
The optional 4–20 mA Interface Board (P/N 3015-3133) (Page 7) allows the operator to take advantage of the  
current loop interfaces provided for connection of external devices to the AGM300. The interface board plugs  
into the socket provided on the main circuit board, and the external devices are then cabled to the AGM300 as  
described on Page 16.  
Instruction 3015-4275  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 - Logic Diagram  
66  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
RS-485 Communication Protocol  
Overview  
The following instructions are intended as a guide for integrating the AGM300 network into a Building  
Management System. If you are unfamiliar with complex systems of this type, it is recommended that you  
contact Bacharach for technical assistance.  
MODBUS RTU Protocol  
The AGM300 monitor communicates with master devices (such as the ADM800 or a Building Management  
System) over the RS-485 serial interface. The monitor is configured with a two wire RS-485 serial bus, and  
therefore, data transmission occurs in ‘half-duplex’ mode. Consequently, only one device may be in transmit  
mode at any given time.  
Programmers should refer to the MODBUS Protocol Reference Guide for details and more comprehensive  
instructions. The guide is available on the Modicon web site at:  
Specific communication software is available at: http://www.bacharach-inc.com/downloads.htm  
Instruction 3015-4275  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
AGM300 MODBUS RTU Operation  
Overview  
The AGM300 and ADM800 are equipped to communicate with other equipment using the MODBUS RTU  
protocol. Using this communication channel a MODBUS master device may communicate with up to 15  
AGM300s on a communications network, exchanging measurement information, alarm data, fault data, history  
(logs and trends) and setup information. Additionally, the MODBUS master can control the operating state of  
an AGM300, placing the AGM300 in any of its different operating modes. The network may be configured such  
that the AGM300s are connected directly to the MODBUS master device, or the MODBUS master device may  
communicate with the AGM300s through the ADM800.  
(NOTE: This document will assume that the reader is familiar with the various parameters used to set up the  
AGM300, as well as the different operational modes that the AGM300 may be placed in. If you are unfamiliar  
with this information, please reference the AGM300 operations manual.)  
Protocol Details  
A 2 wire RS-485 bus is used for transmission, therefore communication occurs in a Half-Duplex mode. The  
AGM300 is a slave device and will respond to queries in the MODBUS RTU format from a master device.  
Two MODBUS functions are supported. They are function 03 (read holding register) and function 16 (Preset  
Multiple registers). Please refer to the MODBUS Protocol Reference Guide (available at  
Using the two MODBUS functions, a master device may read, modify and write data and status information to  
any AGM on the network. AGM data is organized into structures (internal to the AGM300) which can be  
accessed by the MODBUS registers defined in this document. A corresponding set of data structures should  
be maintained by the master device. These master device data structures become the destination for  
responses to read queries and sources for preset register commands. When a read holding register query is  
made by the master device the AGM300 responds by sending the contents of the structure referenced by the  
specified register. After the master validates the AGM300 response using the CRC bytes, it must then move  
the data into its matching data structure before individual items may be accessed or modified. Therefore, the  
master data structure should correspond to the AGM300 data structure byte for byte. Note that some data  
structures have been divided into multiple registers due to MODBUS RTU message length constrains. To  
change a setting in the AGM300, the master device first reads the register structure that contains the data item  
to be modified, makes the desired change, then sends the structure back using the preset multiple register  
function. If the transaction is successful, the AGM300 sends the appropriate MODBUS response. It is the  
responsibility of the master device when making modifications to insure that all parameters trasfered fall inside  
the working limits of the AGM300.  
Notes on AGM300 Polling  
After the AGM300s are setup and operating, the master device need only poll each AGM for its status register  
which contains summary data of the AGM’s alarms, faults, and operating state. If exceptions are detected  
through the status register and more details are required, additional registers can be examined. Also if current  
PPM values are required, the PPM register provides access to current PPM values for all zones. The AGM300  
requires a minimum of 20 seconds to complete a gas concentration measurement for a single zone. Therefore,  
it is not necessary to poll the AGM300 more frequently than once every 15-20 seconds, as there will not be any  
new data available/obtained by more frequent polling. In fact, excessive polling will slow the operation of the  
AGM300. Under no circumstances should the AGM300 be continuously polled at rate faster than 500mS, as  
this could result in erroneous readings by the AGM300.  
68  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Notes on Network Topologies  
AGM300s may be connected directly to the MODBUS network or they may be connected to the network  
through an ADM800. In either case, each AGM300 must have a unique node ID. Up to 15 AGM300s can be  
connected directly to the MODBUS network.  
If AGM300s are accessed via an ADM800 connected to a MODBUS network, the ADM800 “BMS enabled”  
parameter must be set equal to “1” via the “ADM SETUP” screen on the ADM800. The same commands and  
registers are used to communicate with the AGM300 directly or through the ADM800. If the communications is  
through the ADM800, it monitors each MODBUS message to determine if the message is intended for one of  
the AGM300s it is connected to. If it is, the ADM800 passes the message through to the AGM300s. If it is not,  
the message is not passed through. The ADM800 does not make any modifications to MODBUS messages. It  
simply passes the query through to the AGM300, and passes the response back to the MODBUS master. In  
other words, it allows the AGM300s to be logically connected to the MODBUS network, when physically they  
are connected to the local ADM800 network. It is very important to understand that the ADM800 will only pass  
messages through to the AGM300 when the ADM800 is either in the “SYSTEM” screen or the “ZONE VIEW”  
screen. If the ADM800 is in any other screen, it will return a MODBUS “busy” exception response (exception  
code 06).  
All ADM800 screens except the SYSTEM and ZONE VIEW screens have a 10 minute timeout interval, after  
which the screen will return to either the SYSTEM or ZONE VIEW screen, depending on which one it was last  
in. Also, the ADM800 can be password protected such that a password entry is required in order to view  
screens other than the SYSTEM or ZONE VIEW screen.  
Key Comm Protocol Parameters  
MODBUS Mode: RTU only  
AGM300 Baud Rate: Default is 19,200. Programmable as defined in System data register  
Parity: No Parity  
Stop Bits: Default is 1. Can be set for 2 via System data register  
Maximum Response Time: 4000mS when directly accessing the AGM300. 8,000mS when accessing the  
AGM300 through the ADM800.  
Error Checking: CRC per MODBUS specifications  
Also of note is the fact that all data sent out from the AGM-300 is in “little endian” byte order (Least significant  
byte followed by most significant byte). This should be taken into account if the master that process the data is  
a “big endian” type. Non-data information (starting address, number of points, etc.) follows normal MODBUS  
protocol, which is Big Endian.  
Instruction 3015-4275  
69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Summary of Registers  
Register Name Number  
Type Description  
System Data  
Status  
0x0010h  
R/W  
R/W  
AGM System Setup Data  
0x0011h  
Operating summary of faults, alarms  
and status  
Zone Data  
0x12xxh  
R/W  
Setup data for up to 16 zones (xx  
defines zone number)  
CAL Data  
Date/Time  
Sensor Data  
Rel. Hold  
0x0014h  
0x0015h  
0x0016h  
0x0017h  
0x0018h  
0x19xxh  
R/W  
R/W  
R
Cal Factors for all gases  
Set AGM300 date & time  
Raw measurement of sensors  
Release AGM300 out of hold mode  
Put AGM300 into hold mode  
20 most recent fault events (xx = 00  
or 01)  
W
Hold Zone  
Fault Log  
W
R
Alarm Log  
0x1Axxh  
R
20 most recent alarm events (xx = 00,  
01, or 02)  
Serv. Mode  
Rel Serv.  
PPM  
0x001Bh  
0x001Ch  
0x001Eh  
0x3yxxh  
W
W
R
Puts AGM300 into service mode  
Release AGM300 from service mode  
PPM values for all zones  
Zone Log  
R
Trend data for each AGM300 zone (y  
= zone number (starting at 0), xx = 00 – 06)Data  
Data Type Abbreviations  
C ......................Character  
Float.................Floating Point  
I........................Integer  
TIM...................Time  
UC....................Unsigned Character  
UI .....................Unsigned Integer  
70  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
System Data Register  
Register 0x0010h  
R/W  
54 bytes  
Variable  
Type  
Length  
Description  
Type  
UI  
2 bytes  
Indicates EEPPROM has been initialized if value = 300  
DO NOT MODIFY  
REV  
SN  
Float  
UI  
4 bytes  
2 bytes  
1 byte  
Firmware Rev Level DO NOT MODIFY  
Firmware Serial Number DO NOT MODIFY  
Node  
UC  
Network Slave Node # (valid values are 1-15). The default  
is that indicated by the Node DIP Switch on main board.  
Location  
C
C
13 bytes  
1 byte  
Array defining text name of unit  
Stop_Bits  
Number of stop bits used in the AGM300 data stream.  
Default = 1. Other available value is 2  
Aud_Alarm  
UC  
1 byte  
Sounds internal board buzzer on condition; 0 = no buzzer;  
1= Alarm; 2=Fault; 3=Leak; 4=Spill; 5=Evacuate; 6=Monitor  
Off line (DEFAULT = 0)  
Alarm_Ack_  
Mode  
UC  
UC  
1 byte  
1 byte  
Defines Alarm Operation. Manual Acknowledge = 0; Auto  
Acknowledge = 1 (DEFAULT = 0)  
Num_Zones  
Number of install zones (Value initialized during auto  
detect during Power On Self Test).  
UNUSED  
TIM  
UC  
13 bytes  
1 byte  
UNUSED  
Rezero_Mode  
Defines rezero mode. Auto Rezero = 0; Rezero every  
zone = 1 (DEFAULT = 0)  
Hold_Time  
Det_Limit  
UI  
2 byte  
1 byte  
Length of zone hold interval in minutes (DEFAULT = 15  
minutes)  
UC  
Minimum detection limit (in PPM). Concentrations less  
than or equal this value will read as 0 PPM (DEFUALT = 0  
PPM)  
Avg_Size  
UC  
Float  
UI  
1 byte  
Size of running average used in computing PPM value. DO  
NOT MODIFY.  
Loop2_factor  
Serv_Mode_TO  
RS-485_BAUD  
4 bytes  
2 bytes  
2 bytes  
Defines PPM current loop output. (DEFAULT =  
0.16mA/PPM)  
Service Mode Timeout value (in minutes). (DEFAULT = 60  
MINUTES)  
UI  
BAUD RATE for RS-485 connection (between ADM800  
and AGM300 or MODBUS master and AGM300 depending  
on the Network topology). Default=9 (19.2K); other values  
are 8=9600, 7=4800  
UNUSED  
UI  
2 bytes  
UNUSED  
Instruction 3015-4275  
71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Status Register  
Register 0x0011h  
R/W  
10 bytes  
Variable  
Type  
Length  
Description  
Mode  
UC  
1 byte  
Defines Operating Mode of AGM300. 0 = normal Mode; 1  
= Zone_Hold Mode; 2 = Diagnostic Mode; 3 = Service  
mode. DO NOT MODIFY (use zone hold register or  
service mode register to change this parameter)  
State  
UC  
UC  
1 byte  
1 byte  
Defines AGM300 Current State. 0 = Idle; 1 = Sampling; 2  
= Zeroing; 3 = Warm Up, 4 = Pressure Check DO NOT  
MODIFY  
Measuring  
Value = 1 if unit is acquiring detector signal for running avg.  
DO NOT MODIFY  
Active_Zone  
Max _Alarm  
UC  
UC  
1 byte  
1 byte  
Current Zone being checked. 0=zone1, 1=zone2, etc.  
Indicates highest non-acknowledged alarm level DO NOT  
MODIFY  
Alarm_Count  
UC  
1 byte  
Number of alarms that are currently active DO NOT  
MODIFY  
UNUSED  
UC  
UC  
1 byte  
1 byte  
UNUSED  
Loop_Card  
Value = 1 if 4-20mA card has been detected DO NOT  
MODIFY  
Fault  
UI  
2 bytes  
See Note Below  
Fault Flag Structure uses bitwise access to 16 bit word as defined below:  
Bit 15 (MSB) Clipping Fault  
A/D out of range  
Bit 14  
Bit 13  
Bit 12  
Bit 11  
Bit 10  
Bit 9  
Bit 8  
Bit 7  
Bit 6  
Bit 5  
Bit 4  
Bit 3  
Bit 2  
Bit 1  
Bit 0  
Zero Fault  
No Flow  
Zero voltage outside factory limits  
No flow on any zone  
Purge Flow  
Zone Flow  
Trigger Fault  
Unused  
No flow on purge  
No flow on a particular zone(s)  
IR Source clock trigger missing  
Unused  
Config. Fault  
Unused  
No Zones Enabled  
Unused  
Loop Fault  
RS-485 Fault  
Open Current loop  
Comm Error  
Pressure Fault Pressure out of normal operating range  
Bench Temp Fault  
Bench temperature out of normal operating range  
(LSB) Box Temp FaultBox temperature fault  
72  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zone Data  
Register 0x12xxh R/W 78 bytes  
Each zone for an AGM300 has a separate Zone data structure that is 78 bytes long. The zone number is the  
low order byte in the register address (i.e., Zone 1 data register = 0x1201h)  
Variable  
Location  
Flow OK  
Type  
C
Length  
13 bytes  
1 byte  
Description  
13 byte array, Alpha Numeric Description or Name of Zone  
UC  
Status of Flow check. Value of 1 indicates flow check is  
good. DO NOT MODIFY  
Refrig. Type  
Distance  
UC  
UI  
1 byte  
See note 1 Below (DEFAULT = R134a)  
2 bytes  
2 bytes  
4 bytes  
4 bytes  
1 bytes  
Zone Tubing Length (in feet) (DEFAULT = 100 feet)  
Avg temp at zone (degrees C) (DEFAULT = 25°C)  
Last Measured concentrations (uM/L) DO NOT MODIFY  
Last Measured concentration (PPM) DO NOT MODIFY  
Zone Temp  
Concentration  
Concentration2  
Alarm Ack  
I
Float  
Float  
UC  
Set value to 1 to acknowledge Alarm. NOTE: AGM300 will  
reset this byte to 0 when the Alarm byte (below) is = 0 and  
zone in alarm is sampled. If the alarm condition/byte  
increases (leak>>spill or spill>>evac) the AGM300 will also  
reset this byte to 0  
Alarm  
UC  
1 bytes  
Alarm Status; 0 = no alarm, 1 = leak; 2 = spill;  
evac.  
3 =  
Leak Level  
Spill Level  
Evac Level  
UI  
UI  
UI  
2 bytes  
2 bytes  
2 bytes  
Level to trigger a leak alarm (in PPM) (DEFAULT = 100)  
Level to trigger a spill alarm (in PPM) (DEFAULT = 300)  
Level to trigger a evacuate alarm (in PPM) (DEFAULT =  
500)  
Peak PPM  
Peak Time  
Alarm Time  
Not Used  
UI  
2 bytes  
Highest Recorded PPM in zone  
TIM  
TIM  
13 bytes  
13 bytes  
13 bytes  
2 bytes  
Date an time of highest peak (see note 2 for format)  
Date and time of last alarm (see note 2 for format)  
Log Interval  
UI  
Number of minutes between Log entries (DEFAULT =  
1440)  
NOTE 1: Refrigerant Types Defined from 00h to 20h in the following order : 00h=R11, 01h=R12, 02h=R22,  
03h=R23, 04h=R113, 05h=R114, 06h=R123, 07h=R124, 08h=R134a, 09h=R401a, 0Ah=R402a, 0Bh=R402b,  
0Ch=R404a, 0Dh=R407a, 0Eh=R407c, 0Fh=R409a, 10h=R410a, 11h=R500, 12h=R502, 13h=R503,  
14h=R507, 15h=R508b, 16h=H1301, 17h=R408a, 18h=Future2, 19h=Future3, 1Ah=Future4, 1Bh=Future5,  
1Ch=Future6, 1Dh=Future7, 1Eh=New1, 1Fh=New2, 20h=New3  
NOTE 2: Time Structure Format consists of 13 unsigned character types. They are 1 second digit, 10 second  
digit, 1 minute digit, 10 minute digit, 1 hour digit, 10 hour digit, 1 day digit, 10 day digit, 1 month digit, 10 month  
digit, 1 year digit, 10 year digit, last byte is unused.  
Instruction 3015-4275  
73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Notes on Alarms and Alarm Acknowledge  
The AGM300 can be operated in two different alarm acknowledge modes, Auto and Manual (set via the  
alarm_ack_mode variable in the system data register). For purposes of this discussion, the term “Alarm” refers  
to an AGM300 state where the alarm light is on and the appropriate alarm relay is activated. The term “Alarm  
condition” refers to the external condition (i.e., refrigerant leak) that initially causes the AGM300 to go into an  
alarm. If an alarm occurs it can be handled in one of 3 ways.  
1. Non-Latching Mode. This mode is enabled by setting the AUTO_ACK_MODE parameter in the system  
register to “1”. In this mode, if an alarm condition occurs an AGM300 alarm will be created. If the alarm  
condition is subsequently removed, the AGM300 alarm will automatically be cleared by the AGM300 when  
the zone in alarm is sampled “clear”. Note, in this mode of operation, it is possible for an alarm to occur  
and be cleared without user or MODBUS master intervention. If this is the case, the only evidence of the  
alarm would be contained in the AGM300 alarm log.  
2. Latching Mode with Silence. This mode is enabled by setting the AUTO_ACK_MODE in the system  
register to “0”. In this mode, if an alarm condition occurs, an AGM300 alarm will be created. In order for  
the alarm to be removed the MODBUS master will write a “0” to the ALARM parameter in the ZONE  
register. This will cause the alarm to be “silenced” in the AGM300 (i.e., the alarm relays will return to their  
normal state and the ALARM lamp will be extinguished). The next time the zone with the alarm condition is  
sampled, if the alarm condition still exists, the alarm will be reactivated and the alarm parameter will be  
reset to “1” in the AGM300. Otherwise, if the alarm condition has cleared, no further action is required and  
normal operation will resume.  
3. Latching Mode without Silence. This mode is enabled by setting the AUTO_ACK_MODE in the system  
register to “0”. In this mode, if an alarm condition occurs, an AGM300 alarm will be created. The AGM300  
MODBUS master will then write a “1” to the ALARM ACK parameter in the zone register. The alarm will  
continue to persists (i.e., Relays in alarm state and Alarm light on) until the offending zone is sampled and  
no alarm condition is detected. At that point, the ALARM ACK parameter is automatically cleared by the  
AGM300, as is the ALARM parameter. NOTE: If the ALARM ACK parameter is set to “1” and the ALARM  
CONDITION is upgraded (from leak to spill, or spill to evacuate) the ALARM ACK parameter will  
automatically be cleared to “0” by the AGM300.  
74  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Cal Data Register  
Register 0x0014h  
R/W  
174 bytes  
Variable  
Type  
Length  
Description  
Factor  
Float  
132 bytes  
33 element array containing cal factors for each of the 33  
gases. Order of Cal Factor is defined in Note 1 under Zone  
Data  
Curve  
UC  
C
3 bytes  
Existing Master Curve to use for user defined “NEW GAS”  
New Gas Name  
39 bytes  
3 x 13 array that contains the names for each of 3 new  
gases  
NOTE: “Future” Gases are gases that will be added to the library at a later date. They will require generation  
of new cal curves that will be added to the code. “New” Gases can be added in the field by selecting a similar  
existing cal curve and then defining a cal factor that will bring the accuracy within the desired range. It is  
strongly recommended that any New Gas curves that are added be done in conjunction with help from  
Bacharach, Inc.  
Date Time Register  
Register 0x0015h  
R/W  
14 bytes  
Variable  
Type  
Length  
Description  
Date_Time  
Time  
14 bytes  
Contains current time and date. Structure is defined as in  
note 2 of zone data  
Time Structure Format consists of 14 unsigned character types. They are 1 second digit, 10 second digit, 1  
minute digit, 10 minute digit, 1 hour digit, 10 hour digit, a day digit, 10 day digit, 1 month digit, 10 month digit, 1  
year digit, 10 year digit, day of the week, last byte is unused.  
Instruction 3015-4275  
75  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Sensor Data Register  
Register 0x0016h  
R
82 bytes  
Variable  
Pressure  
P_Volts  
Type  
Length  
Description  
Float  
Float  
Float  
Float  
Float  
Float  
Float  
Float  
Float  
Float  
UI  
4
4
4
4
4
4
4
4
4
4
2
4
4
4
4
4
4
4
Manifold Pressure is PSIA  
Pressure sensor output Voltage  
Vacuum_P  
Ambient_P  
Box_T  
Pressure with all value closed and pump on in PSIA  
Absolute Ambient Pressure in PSIA  
Enclosure Temperature in Degrees C  
Box temp sensor output voltage  
Box_T_Volts  
Bench_T  
Bench_T_Volts  
Bench_Z_T  
Box_Z_T  
PkPk_int  
PkPk  
Optical bench temperature in Degrees C  
Bench temp sensor output voltage  
Optical bench temp in degrees C at last zero interval  
Box temp in degrees C at last zeroing  
Current peak to peak A/D counts from detector  
Current peak to peak voltage from detector  
Running average voltage from detector  
Voltage measured at last zeroing  
Float  
Float  
Float  
Float  
Float  
Float  
Float  
Ave_PkPk  
Zero_PkPk  
Noise  
Largest Change in running average  
Current absorbance value  
AU  
Ave_AU  
Running Average of absorbance level  
Bench_PPM  
PPM in bench based on zone gas selection (uncorrected  
for pressure and temperature)  
STP_PPM  
Float  
Float  
Float  
4
4
4
PPM corrected to STP (1 atm, 25 deg C)  
Micromoles/liter in bench (uncorrected0  
Micromoles/liter corrected to ambient pressure  
Bench_UML  
Ambient_UML  
76  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Release Zone Hold Register  
Register 0x0017h  
Description  
W
10 bytes  
Variable  
Type  
Length  
Rel_Hold  
*
*
See description of STATUS REGISTER  
Hold Zone Register  
Register 0x0018h  
Description  
W
10 bytes  
Variable  
Type  
Length  
Zone_Hold  
*
*
See description of STATUS REGISTER  
Notes Concerning the AGM300 Hold Mode  
The AGM300 can be made to hold or “dwell” on a particular zone if necessary. The length of the hold time is  
defined by the HOLD TIME parameter in the System Data Register. In order to place the AGM300 into hold  
mode use the following sequence:  
Read the AGM300 Status Register (0x0011h)  
Modify the content of the status register structure to change the MODE parameter to zone hold mode and  
the active zone parameter to the zone which you wish to hold  
Send this updated status register structure back to the AGM300 using PRESET MULTIPLE REGISTER  
COMMAND to the HOLD ZONE REGISTER (0x0018h).  
To Release the Zone Hold Use the Following Sequence of Steps  
Read the AGM300 Status Register (0x0011h)  
Modify the content of the status register to change the MODE parameter to normal mode and the active  
zone parameter to the zone which you would like to resume normal activity on  
Send this updated status register structure back to the AGM300 using PRESET MULTIPLE REGISTER  
COMMAND to the RELEASE HOLD REGISTER (0x0017h).  
Instruction 3015-4275  
77  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fault Log Register  
Register 0x1900, 0x1901h  
R
302 bytes  
These registers contain the 20 most recent fault events, the time they occurred, and a pointer to the most  
recent event. The data is split into 2 registers. The first register contains 200 bytes and the second register  
contains 102 bytes. The results of these two register reads should be recombined into the Fault Log Data  
Structure after both have been received  
Variable  
Type  
Length  
Description  
Fault  
UI  
40 bytes  
20 most recent fault events. Each event is decoded as  
indicated in Fault Flag Structure given after the Status  
Register Description  
Time  
TIM  
260 bytes  
Time of each fault occurrence. TIM value as defined in  
NOTE 2 of Zone Data  
Ptr  
UC  
UC  
1 byte  
1 byte  
Pointer to most recent event  
Unused  
Alarm Log Register  
Register 0x1A00h, 0x1A01h, 0x1A02h R 582 bytes  
These registers contain the 20 most recent alarm events, the time they occurred, and a pointer to the most  
recent event. The data is split into 3 registers and should be recombined into an appropriate structure after all  
three registers have been received. Register 0x1A00h contain 200 bytes, Register 0x1A01h contains 200  
bytes, and register 0x1A02h contains 181 bytes.  
Variable  
Type  
Length  
Description  
Event  
UC  
320 bytes  
20 most recent alarm events. Each event contains 1 byte  
for each zone. Each zone Byte is defined as 0=No Alarm,  
1=Leak Alarm, 2=Spill Alarm, 3=Evac Alarm.  
Time  
TIM  
260 bytes  
Time of each alarm event. TIM value as defined in NOTE 2  
of Zone Data  
Ptr  
UC  
UC  
1 byte  
1 byte  
Pointer to most recent event  
Unused  
78  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Mode Register  
Register 0x001Bh  
W
10 bytes  
Variable  
Type  
Length  
Description  
See description of STATUS REGISTER  
Rel_Svc_Mode  
*
*
Release Service Mode Register  
Register 0x001Ch  
Description  
W
10 bytes  
Variable  
Type  
Length  
Ent Svc_Mode  
*
*
See description of STATUS REGISTER  
Notes Concerning the AGM300 Service Mode  
The AGM300 can be placed into service mode if necessary. During service mode the unit will take no  
measurements, any and all alarms are silenced, and all relays are opened. The unit automatically comes out  
of service mode after a preset interval defined by the service_mode_TO parameter in the System Data  
Register. In order to place the AGM300 into service mode use the following sequence:  
Read the AGM300 Status Register (0x0011h)  
Modify the content of the status register structure to change the MODE parameter to service mode.  
Send this updated status register structure back to the AGM300 using PRESET MULTIPLE REGISTER  
COMMAND to the SERVICE MODE REGISTER (0x001Bh).  
To Release the Unit from Service Mode Use the  
Following Sequence of Steps  
Read the AGM300 Status Register (0x0011h)  
Modify the content of the status register to change the MODE parameter to normal mode.  
Send this updated status register structure back to the AGM300 using PRESET MULTIPLE REGISTER  
COMMAND to the RELEASE SERVICE MODE REGISTER (0x001Ch).  
Instruction 3015-4275  
79  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PPM Register  
Register 0x001Eh  
R
32 bytes  
Variable  
Type  
UI  
Length  
Description  
PPM  
32 bytes  
16 Unsigned Integers that represent the PPM values for  
each AGM300 zone  
NOTE: 16 values are returned independent of the number of actual zones installed in the unit. The master  
device is required to know how many zones are installed in the unit (available in the System Register) in order  
to properly interpret the data.  
Zone Log Registers  
Register 0x3xyyh  
R
1502 bytes  
These registers are used to transfer the zone log data. Each zone has a circular log of 100 past data points.  
The period between data points is defined by the Log Interval parameter in each corresponding Zone Data  
Register. The data for each zone is defined by the “x” place in the above register address. For zone 1 the  
Register address is 0x30yyh, for zone 2 the register address is 0x31yyh, etc. The data for each zone is sent in  
8 consecutive registers due to MODBUS RTU message length constrains. The addresses are defined by the  
“yy” place in the above address. For zone 1, all log data can be obtained by reading 0x3000h, 0x3001h,  
0x3002h, …… , 0x3007h. The first seven registers contain 200 bytes each and the last register contains 102  
bytes. After all registers have been received the data should be reassembled into the full data structure.  
Variable  
Index  
Type  
UI  
Length  
2
Description  
Point to current reading  
Time  
TIM  
1300  
Time record for each of the 100 log points. The format  
for the TIM type is defined in note 2 of zone data  
PPM  
UI  
200  
Last 100 log points (2 byes per point)  
MODBUS EXCEPTION RESPONSES  
The following MODBUS exception response are supported by the AGM300:  
01  
02  
06  
Illegal Function  
Illegal Data Address  
Slave Device Busy (Occurs only when AGM300 is connected to the bus through an ADM800 and the  
ADM800 is not in the SYSTEM or ZONE VIEW screen)  
80  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARRANTY and SERVICE  
Warranty  
Subject to the terms herein, Seller warrants the original equipment cataloged or manufactured by the Seller and  
furnished hereunder is free from defects in material and workmanship, and will be of the kind and quality  
designated or described on the quotation. The Seller will warrant the original equipment for 24 months from the  
date of shipment from the manufacturing facility unless different or additional conditions are specified and shall  
be binding through a written contract signed by an officer of the Seller.  
Seller also warrants that certain consumable parts as hereunder identified have a one (1) year warranty period  
from the date of shipment and will be free from all defects in material during that period. Seller does not  
warrant these parts however, beyond the initial installation due to variables associated with normal usage.  
These certain consumable parts include, but are not limited to, the air flow filters.  
If within 24 months of the date of shipment, the equipment does not meet the warranties specified above, and  
Purchaser notified Seller of this promptly, Seller shall there upon correct any such defects or any non-  
conformance to the specifications within a reasonable time period. Material found to be defective shall be  
returned to the seller freight prepaid. At the Seller’s option, the defective material shall be repaired or replaced  
and returned to the purchaser via the lowest freight rate available. Any repairs or alternations made by  
unauthorized personnel will automatically void warranty.  
Labor cost for warranty service at the manufacturing facility is the Seller’s responsibility for a period of 90 days  
following date of shipment. Beyond 90 days, Purchaser is responsible for all labor costs. Purchaser must  
contact the Seller’s Service Department prior to shipment and obtain a Return Authorization (RA) number. The  
package must be clearly marked with the assigned RA number. Purchaser may elect to have Seller’s Service  
Department repair the equipment at Purchaser’s facility. Purchaser must contact Seller’s Service Department  
to arrange for a Service call. Parts will be free under the above terms, but there will be a charge for labor and  
travel expenses from Seller’s facility in New Kensington, PA..  
The conditions of any tests or equipment hereunder alleged to be defective shall be mutually agreed upon and  
Seller shall be notified of, and may be present or represented at all tests that may be made. Seller’s liability to  
Purchaser or that of Seller’s Agent/Representative, arising out of the supplying of the equipment or its use,  
whether based on warranty, contract, or negligence shall not in any case, exceed the cost of correcting defects  
in the equipment as herein provided but shall, in all events, be limited to a maximum of the purchase price.  
Upon the expiration of the applicable warranty period mentioned above, all such liability shall terminate. The  
foregoing shall constitute Purchaser’s sole remedy and Seller’s sole liability. IN NO EVENT SHALL SELLER  
BE LIABLE FOR SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES.  
Register Your Warranty by Visiting  
Service  
Pursuant to the stated warranty, Bacharach, Inc. warrants the products to be free of manufacturers defect for  
two (2) years from the date of shipment or installation and will replace any part free of charge during that period.  
Labor beyond ninety (90) days of shipment or installation is subject to charge. FAILURE TO LOCATE THE  
AGM300 OR ADM800 IN A LOCATION COMPATIBLE WITH THE CRITERIA SET FORTH IN THIS MANUAL  
CONSTITUTE ABUSE OF THE EQUIPMENT AND MAY VOID THE WARRANTY. Shipments are tracked by  
the factory and the purchaser will be responsible for providing acceptable documentation to prove the  
installation date. FAILURE TO NOTIFY THE FACTORY WITHIN TWO (2) WEEKS OF INSTALLATION OF  
THE EQUIPMENT WILL VOID THE COVERAGE OF THE LABOR SPECIFIED IN THE WARRANTY. Should  
any of the products covered (refer to following section - Products Covered under Warranty and Service Policy)  
incur a failure or damage contact the factory service department (724-334-5051) and request instructions.  
Instruction 3015-4275  
81  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Return Procedure  
To return equipment to the factory for repair a Return Authorization (RA) number should be requested from  
the factory. Equipment will not be received without an RA number. Returned equipment should be sent to  
the factory at the following address:  
Bacharach, Inc.  
621 Hunt Valley Circle  
New Kensington, PA 15068  
Attn.: Service Department  
Field Repair  
Field repair of equipment will be considered after all reasonable attempts to diagnose and rectify the  
problem via phone or fax have been exhausted or if return of the equipment to the factory is not practical.  
To initiate warranty service in the field, a customer purchase order must be issued, prior to dispatch, for  
the estimated travel time, expenses and applicable service fees for a Bacharach representative to travel to  
the site. Bacharach will bill expenses to the customer at actual cost. The customer will be invoiced for all  
applicable charges for parts, labor and expense at the Non-Warranty rates if it has been determined that  
the failure is a result of abuse of the equipment, improper installation, or failure to thoroughly read and  
comply with the instructions provided in the Operations and /or Service Manuals. Scheduling of all field  
servicing is at the discretion of the Factory Service Department and will be based upon customer need and  
the availability of personnel. Warranty field service shall be limited to the continental United States.  
Bacharach, Inc. is not liable for any charges, from the customer, purchaser or any third party that may be  
acting on the behalf of the purchaser, customer or end user, resulting from non-warranty field service, field  
upgrades, preventative maintenance, etc. Bacharach, Inc. is not liable for any charges, from the customer,  
purchaser or any third party that may be acting on the behalf of the purchaser, customer or end user,  
resulting from warranty field service or repair activities unless prior approval is granted by an authorized  
member of Bacharach, Inc. prior the dispatch and performance of the field service or repair activity. This is  
to include but not limited to: manpower requirements for Bacharach and/or non-Bacharach personnel to  
perform or assist with the field service or repair activity.  
Products Covered under Warranty and Service Policy  
The following items are covered under the stated warranty for Parts and Labor:  
Model  
Description  
AGM300  
ADM800  
Ammonia Gas Monitor  
Ammonia Display Module  
82  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
AGM300 - Specifications  
Product Type:  
Ammonia gas, multiple area infrared monitoring system for low level continuous  
monitoring of ammonia refrigerant gas used in most commercial refrigeration systems.  
System design supports compliance to the refrigerant monitoring requirements of  
ANS/BSR ASHRE 15-1994.  
Coverage:  
4-point standard, expandable to 16 points in 4 point increments  
Detector Type:  
Gas:  
Infrared non-dispersive, computer enhanced  
Ammonia, NH3, R-717  
1 ppm  
Sensitivity:  
Measuring Range:  
Accuracy:  
25 to 10,000 ppm  
±10 % of reading  
Front Panel:  
3 Indicator lights:  
One green light will glow steady when unit is operating normally, or flashing when  
unit is in warm-up mode  
One red light will glow steady when any point is above the alarm setting, or will be  
flashing when the unit is in service mode  
One yellow light will glow when there is a system fault  
Temperature Drift:  
Sampling Mode:  
Re-Zero:  
±0.3% of reading per degrees C  
Automatic or manual (hold)  
Auto or on zone change  
Response Time:  
System Noise:  
5 to 120 seconds - depending on air line length and number of zones  
Less than 40dB(A) @ 10 feet  
Monitoring Distance: Up to 500 feet maximum air line length (each zone)  
Conditioned Signal: Dual optional 4–20 mA isolated outputs; Channel 1 = zone area; Channel 2 = PPM  
Alarms:  
Four SPDT alarm contacts are provided (Rated 5 amp 120 VAC)  
Three assigned to PPM level alarms - One assigned to system faults  
Communications:  
Full two-way communication with ADM800 Ammonia Display Module or Building  
Management System via RS-485 serial interface.  
RS-232C communication port standard  
MODBUS RTU communication protocol standard  
Power Safety Mode: Fully automatic system reset. All programmed parameters retained.  
Operating Temp:  
Ambient Humidity:  
Size/Weight:  
AC Power:  
32 to 113 °F (0 to 45 °C)  
5 to 90% (non-condensing)  
17"H x 12"W x 5.5"D - 15 lbs.  
120 or 230 VAC, 50/60 Hz, 21 Watts  
UL #61010-1 and CE Mark  
Two (2) Years from date of shipment  
2,000 Meters  
Certification:  
Warranty:  
Altitude Limit:  
Instruction 3015-4275  
83  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ADM800 - Specifications  
Product Description: The ADM800 Ammonia Display Module provides remote programming, interrogation  
and display functionality to support the AGM300 Ammonia Gas Monitor. The system  
design supports compliance to the refrigerant monitoring requirements of ANS/BSR  
ASHRE 15-1994.  
Inputs:  
The ADM800 accepts inputs from up to four AGM300 monitors. It offers a wide variety  
of displays and can fully program any associated monitor.  
Display:  
Back lit LCD  
Power Indication:  
Alarm Indication:  
Fault Indication:  
Green LED glows when AC power is applied  
Red LED glows and alarm condition is displayed  
Yellow LED glows and fault condition is displayed  
Password Protection: The ADM800 can be password protected to prevent the unauthorized editing of setup  
parameters without disturbing the ability to monitor system outputs  
Power Safety Mode: Fully automatic system reset; all parameters maintained  
Monitoring Distance: Up to 4500 feet from furthest AGM300  
Communications:  
Full two-way communication with AGM300 monitors via RS-485 interface. A second  
RS-485 interface provided for connection to a Building Management System.  
Relays:  
Two SDDT programmable alarm contacts are provided (Rated 5 amp 120 VAC)  
32 to 113 °F (0 to 45 °C)  
Operating Temp:  
Ambient Humidity:  
Size/Weight:  
AC Power:  
5 to 90% (non-condensing)  
11"H x 10"W x 3"D - 5 lbs.  
120 to 230 VAC, 50/60 Hz, 21 Watts  
Wall mount  
Mounting:  
Certification:  
Warranty:  
UL 61010-1 and CE Mark  
Two (2) Years from date of shipment  
2,000 Meters  
Altitude Limit:  
84  
Instruction 3015-4275  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Notes:  
Instruction 3015-4256  
85  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Headquarters:  
621 Hunt Valley Circle, New Kensington, PA 15068  
Ph: 724-334-5000 • Fax: 724-334-5001 • Toll Free: 1-800-736-4666  
Printed in U.S.A.  
® Registered Trademark of Bacharach Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Actiontec electronic Network Router GT701D User Manual
Acu Rite Weather Radio 01098CDI User Manual
AEG Cooktop 5033 V User Manual
Aiwa Portable CD Player CDC X30MP User Manual
AKG Acoustics Microphone D 58E User Manual
Amana Refrigerator ARS2661BB User Manual
Axis Communications Security Camera P5532 User Manual
Barco Projection Television PGWU 61B User Manual
Beckett Water Gardening Outdoor Fountain LT20 User Manual
Beko Ventilation Hood CWB 7441 X User Manual