Dear ACCU-CHEKSpirit insulin pump owner,  
Thank you for choosing the ACCU-CHEK Spirit insulin pump.  
This guide will help you to get the most from your ACCU-CHEK Spirit insulin  
pump. Even if you have used an insulin pump before, it’s important that you  
do these things before you use the ACCU-CHEK Spirit insulin pump:  
̈ read this guide carefully  
̈ work with your doctor or healthcare team to understand how to use  
the pump  
̈ test your blood glucose level often to find your personal settings and  
to avoid getting the wrong dosage of insulin. (An ACCU-CHEK blood  
glucose meter is one way to test blood glucose levels; talk to your  
doctor or healthcare team to decide the best way for you.)  
It’s a good idea to inform family members or friends about diabetes and pump  
therapy so that they can help you in an emergency.  
Disetronic Medical Systems is proud to help you manage your diabetes therapy.  
However, Disetronic Medical Systems cannot recommend specific diabetes  
therapy. Always follow the instructions given to you by your doctor or healthcare  
team. If you have questions about your pump, or problems with it, refer to this  
guide and talk to your doctor or healthcare team. We also invite you to contact  
Pump Support at 1-800-688-4578.  
Note:  
Your insulin pump is a valuable medical device. Consider adding your  
pump to your homeowner’s insurance policy to protect you in case it is  
damaged or lost. Ask your insurance agent for details.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
How to use this guide  
This guide is for version 2.XX of the ACCU-CHEK Spirit insulin pump.  
Note:  
This user guide presents you with sample screens. The screens in this  
user guide may look slightly different from your pump screens. If you  
have questions about what you see on your pump, contact Pump  
Support at 1-800-688-4578.  
When you become familiar with your pump, you can quickly refer to the back  
of this user guide for tables and definitions that can help you to understand  
pump alerts and warnings, and the symbols on the pump screen. There is also  
a glossary at the back that defines terms used in this guide, and a table that  
lists the technical specifications for your pump. All of this information can help  
you and your doctor or healthcare team to understand the pump and use it to  
best manage your insulin pump therapy.  
Warnings and Cautions  
Read all warnings and cautions in this guide before you use the pump. This  
information is important to protect your health and the function of your pump.  
If you have any questions about a warning or caution, contact Pump Support  
at 1-800-688-4578. Warnings and Cautions are indicated by triangles.  
WARNINGS are indicated by a triangle and the word  
WARNING.  
!
L
WARNING  
CAUTIONS are indicated by a triangle and the word  
CAUTION.  
!
L
CAUTION  
2
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Working safely with your ACCU-CHEK Spirit insulin pump  
Your ACCU-CHEK Spirit insulin pump meets the safety requirements of national  
medical device legislation and meets or exceeds the international standards for  
Electromagnetic Compatibility regarding the pump’s use. This section introduces  
you to the intended use of the ACCU-CHEK Spirit insulin pump, to its safety  
features, and to warnings and safety information you need to know while using  
your pump.  
While your pump is in use, two internal safety systems monitor and control the  
pump’s functioning. The pump performs more than nine million safety checks  
every day. If your pump detects a change from its normal state, an alert or an  
error message is delivered.  
See the sections “Alerts and Errors” (page 147) and “Appendix A: Technical Data”  
(page 165) for information on how to address errors and alerts.  
Intended use of your ACCU-CHEK Spirit insulin pump  
The ACCU-CHEK Spirit insulin pump is intended for the subcutaneous continuous  
delivery of insulin, at set and variable rates, for the management of diabetes  
mellitus in persons requiring insulin as prescribed by a physician.  
The ACCU-CHEK Spirit insulin pump is intended for the use of U100 rapid-  
acting insulin.  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3
Guarantee  
Any changes or modifications to the ACCU-CHEK Spirit insulin pump that  
are not expressly approved by Disetronic Medical Systems, Inc. could void  
the pump warranty. Do not try to fix the device yourself.  
4
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Chapter 1: Understanding Your New System _____________________9  
The ACCU-CHEK Spirit Insulin Pump __________________________9  
Display ________________________________________________9  
Keys and Key Combinations _______________________________11  
STATUS Screen Messages, Beeps, and Vibrations_______________15  
Accessories and Disposables ______________________________17  
Emergency Kit __________________________________________23  
Chapter 2: Getting Started __________________________________25  
Inserting (or Changing) the Battery__________________________25  
System Check Start-Up Process ____________________________28  
Setting Time and Date____________________________________32  
Your Personal Settings ___________________________________35  
Your Basal Rate Profile ___________________________________36  
Programming Your Basal Rate Profile________________________37  
Preparing the Cartridge___________________________________42  
Connecting the Cartridge, Adapter, and Infusion Set_____________47  
Inserting the Cartridge ___________________________________49  
Priming the Infusion Set __________________________________53  
Preparing the Infusion Site and Inserting the Infusion Set ________56  
Chapter 3: Using Your Pump_________________________________59  
Before Using ___________________________________________59  
Wearing Your Pump______________________________________59  
When Not to Wear Your Pump ______________________________61  
Your Insulin Pump and Water ______________________________63  
Starting Insulin Delivery __________________________________65  
Stopping Insulin Delivery _________________________________66  
Interruption of Insulin Pump Therapy ________________________68  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5
Changing the Infusion Set_________________________________70  
Changing the Cartridge and Infusion Set _____________________73  
Chapter 4: Boluses and Temporary Basal Rates__________________75  
Understanding Boluses ___________________________________75  
Programming a Standard Bolus ____________________________77  
Programming an Extended Bolus ___________________________83  
Programming a MultiWave Bolus ___________________________88  
Programming a Temporary Basal Rate (TBR) __________________93  
Chapter 5: User Menus _____________________________________99  
Selecting a User Menu __________________________________100  
STANDARD User Menu___________________________________103  
Data Transfer__________________________________________104  
Setup for Data Transfer __________________________________105  
Reviewing Pump Data ___________________________________107  
SETUP MENU Standard __________________________________113  
ADVANCED User Menu___________________________________122  
Basal Rate Profiles _____________________________________123  
Alarm Clock___________________________________________127  
SETUP MENU Advanced__________________________________129  
CUSTOM User Menu_____________________________________136  
Chapter 6: Care and Maintenance ___________________________137  
Care of Your ACCU-CHEK Spirit Insulin Pump _________________137  
System Inspection Checklist______________________________137  
What to Do When You Drop Your Pump ______________________139  
Storing Your ACCU-CHEK Spirit Insulin Pump _________________140  
Disposal _____________________________________________140  
Maintenance __________________________________________141  
Repair _______________________________________________142  
Troubleshooting________________________________________143  
6
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7: Alerts and Errors________________________________147  
Confirming an Alert or Error ______________________________148  
List of Alerts __________________________________________149  
List of Errors __________________________________________154  
Appendixes_____________________________________________165  
Appendix A: Technical Data_______________________________165  
Appendix B: Configuration Parameters ______________________177  
Appendix C: Abbreviations _______________________________180  
Appendix D: Beeps and Melodies __________________________181  
Appendix E: Vibrations __________________________________182  
Appendix F: Symbols____________________________________183  
Appendix G: Sterile Products and Accessories ________________188  
Appendix H: Glossary ___________________________________190  
Warranty _____________________________________________194  
Index _________________________________________________195  
Diagram of User Menus ___________________________________198  
Quick Reference Table for Alerts ____________________________199  
Quick Reference Table for Errors ____________________________200  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7
This page left blank intentionally.  
8
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 1: Understanding Your New System  
The ACCU-CHEK Spirit Insulin Pump  
Display  
Menu key  
Up key  
Down key  
Check key  
Infusion set  
Adapter  
Cartridge  
Check your insulin pump and its sterile products, like an  
infusion set and accessories, at least every three hours.  
!
L
Immediately inspect your insulin pump if it is dropped.  
Do not use your insulin pump if chips or cracks are visible.  
Chips and cracks may allow water, dust, insulin, or other  
foreign substances to enter your insulin pump and lead  
to malfunction.  
CAUTION  
Display  
Your insulin pump has a Liquid Crystal Display (LCD) screen that displays  
important information. Check the display at least every three hours during  
the day, before you go to sleep, and if you are not able to hear the beeps or  
alerts or feel the vibrations. This is the only way you will be aware of important  
changes in your pump.  
Chapter 1: Understanding Your New System  
ACCU-CHEK Spirit Insulin Pump  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
If you discover incomplete characters, numbers, or symbols  
on the display, put your insulin pump into STOP mode and  
perform a system check by simply removing the battery for a  
few seconds. When you reinsert the battery the system check  
will occur. If the incomplete characters, numbers, or symbols  
remain, your insulin pump may need to be returned since you  
may not receive the correct information for operating your  
pump. Put your insulin pump into STOP mode and contact  
Pump Support at 1-800-688-4578 immediately.  
!
L
CAUTION  
Note:  
To make the display easier to read, you can flip the display orientation  
by 180° and adjust the display contrast lighter or darker. See the sections  
“Display Orientation” (page 120) and “Display Contrast” (page 135) for  
more information.  
Backlight  
The backlight helps you to see the display in poorly lighted areas. You can turn  
on the backlight from the RUN screen, from the STOP screen, or when you are  
browsing the menus. Press the a key to turn on the backlight. The backlight  
turns itself on when an alert or an error message appears. The backlight turns  
itself off after 30 seconds if no keys are pressed.  
Hint:  
Start programming with the a key in areas of poor lighting as this will  
also turn on the backlight.  
Chapter 1: Understanding Your New System  
10  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Keys and Key Combinations  
The pump will beep each time a key is pressed, unless the beep tone is set to 0.  
Never press any of the keys with a sharp or pointed item  
such as the tip of your fingernail. This could damage the  
casing of your insulin pump. Use the pad of your fingertip  
instead to avoid puncturing the keys.  
!
L
CAUTION  
Single Keys  
Each key has a specific function when it is pressed by itself. The following  
table describes the function of each.  
Key Name Function  
d
Menu  
scrolling through menus, functions, and  
information screens  
f
Check  
selecting a menu  
saving changes and exiting screen  
viewing the QUICK INFO screen  
a
s
Up  
increasing a setting  
moving forward in the information screens  
turning on the backlight  
Down  
decreasing a setting  
moving backward in the information screens  
Either a or s can be used for programming or cancelling a Standard  
bolus. Pressing either key for 3 seconds (until you hear a melody) will  
also temporarily turn off the STOP-Warning.  
Chapter 1: Understanding Your New System  
ACCU-CHEK Spirit Insulin Pump  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Note:  
Increase or decrease the values by pressing the a or s keys. If  
you want to make large changes to a value, in most cases you can do  
this faster by pressing and holding down the a or s key (scrolling)  
until the number you want is displayed. If needed, the value can still  
be corrected by a single press of the s or a. Your pump will beep  
once when you start scrolling.  
Key Combinations  
Keys can be pressed in pairs for additional functions. In the following table,  
a plus symbol (+) between 2 keys means that both keys must be pressed  
simultaneously.  
Combination Function  
d + a  
d + s  
a + s  
exiting menus, function screens, and  
information screens without saving  
moving backwards to a previous menu  
unlocking the keys (press and hold both  
keys until you hear three beeps in RUN mode  
and one beep in STOP mode.)  
copying an hourly basal rate to the next hour(s)  
If the keys on your pump are not functioning properly, or you  
are not able to identify them correctly, disconnect your  
insulin pump and contact Pump Support at 1-800-688-4578.  
!
L
WARNING  
Chapter 1: Understanding Your New System  
12  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Locking Keys Using KeyLock Feature  
The KeyLock feature lets you lock all four keys on your pump as a safety measure  
against accidentally activating functions. You may want to use this function  
while sleeping, or during sports, for example. The KeyLock function is turned on  
in the STANDARD SETUP MENU. All instructions in this user guide are given on  
the assumption that the KeyLock is off. For more information see the section  
“Turning KeyLock On or Off” (page 113).  
Screens  
Your pump displays various screens on its display for you to access information  
and functions.  
RUN Screen  
basal rate  
time  
current hourly  
basal rate  
KeyLock locked  
pump timer  
cartridge low  
battery low  
active basal rate profile  
During normal use, your pump is in RUN mode and insulin is delivered into  
your body based on the basal rate you have programmed. The RUN screen is  
the “home” screen and is displayed when the pump is in RUN mode. After you  
program or review information, the pump returns to the RUN screen. This  
screen shows the current time, the current basal rate profile, and the current  
hourly basal rate in units per hour. The RUN screen will also display reminder  
information (such as “cartridge low”) and special functions (such as KeyLock  
locked or unlocked) in the form of symbols.  
See the sections “Starting Insulin Delivery” (page 65) for instructions on starting  
your pump and “Appendix F: Symbols” (page 183) for more information about  
the symbols.  
Chapter 1: Understanding Your New System  
ACCU-CHEK Spirit Insulin Pump  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STOP Screen  
date  
time  
KeyLock unlocked  
pump timer  
cartridge low  
battery low  
The STOP screen shows the current time and date, the STOP symbol, reminder  
information (such as “battery low”), and special functions (such as KeyLock locked  
or unlocked). Some functions must be programmed while your pump is in STOP  
mode, when insulin is not being delivered. Your pump must also be in STOP mode  
before you can change cartridges. Additionally, always disconnect your pump from  
the infusion site and place the pump in STOP mode when you prime the infusion  
set and/or when you transfer data. See the section “Stopping Insulin Delivery”  
(page 66) for instructions on stopping your pump.  
QUICK INFO Screen  
If you press the f key from the RUN or STOP  
screen, your pump displays the remaining cartridge  
content in units. Symbols for activated functions and  
features (like the beep and vibration alarm signals) are displayed. Your pump will  
return to the RUN or STOP screen after 8 seconds.  
Information, Function, and Setup Screens  
You will need to access various screens so you can view stored information,  
program the pump, perform a function, or change a setting. You can reach  
these screens from either the RUN or STOP screen.  
Chapter 1: Understanding Your New System  
14  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Follow the steps below to navigate through the screens.  
1. Press the d key to move to a screen displaying your first choice. Think  
of these screens as menu choices.  
2. Press the d key again to see your next choice on the menu. After you  
reach the last menu choice you will “loop” back to the first menu choice.  
3. When you reach the screen displaying the choice you want (e.g., SET THE  
TIME AND DATE), press f. The next set of screens are called function  
screens. They all relate to the choice you made. There may be only one  
function screen or multiple screens.  
4. Press the a or s key to increase or decrease the selected value. Or,  
press the d key again to move to the next function screen you want to  
change. After you reach the last function screen you will “loop” back to  
the first function screen.  
5. Press f to save your changes. The pump returns to the RUN or STOP  
screen and the changes are saved.  
Note:  
The menu choices that appear depend on the menu that is selected.  
Your pump can display a Standard, Advance, or Custom user menu.  
See the section “User Menus” (page 99) for more information.  
Pressing d+a simultaneously lets you move backwards in the  
menu structure or to return to menus you just skipped. If you are  
within a function screen, pressing d+a simultaneously lets you  
move up to the corresponding menu without saving the current changes.  
STATUS Screen Messages, Beeps, and Vibrations  
Beeps, vibrations, and messages on the display inform you of the status of  
your insulin pump. Beeps and/or vibrations are used as alarm signals. Your  
insulin pump also beeps when a key is pressed unless the beep volume is  
turned off. The beep volume can be turned up or down.  
Chapter 1: Understanding Your New System  
ACCU-CHEK Spirit Insulin Pump  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Check the display of your insulin pump at least every three  
hours during the day and before you go to sleep, especially  
if for any reason you are unable to hear the beeps or feel  
the vibrations. This is the only way you will be notified of  
changes in your insulin pump in a suitable amount of time.  
!
L
WARNING  
Alarm Signals  
Your pump gives alerts and errors using beeps and vibrations. You can turn off  
either the beep or the vibrations but for your safety you cannot turn off both at  
the same time for the initial alarm signal. See the sections “Alerts and Errors”  
(page 147), “Adjusting the Beep Volume” (page 116), and “Setting Up Alarm  
Signals” (page 117) for more information.  
STOP-Warning  
The STOP-Warning is turned on by the pump each time your pump switches from  
RUN mode to STOP mode, and when a battery is inserted. The STOP-Warning  
is a long beep and a vibration occurring every minute. This warns you that  
insulin delivery has been interrupted.  
To Turn Off the STOP-Warning  
Press either a or s for 3 seconds to turn off the STOP-Warning. A melody  
plays indicating the STOP-Warning is turned off.  
Note:  
The volume of the STOP-Warning cannot be changed. It is independent  
of the programmed beep volume.  
Chapter 1: Understanding Your New System  
16  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Accessories and Disposables  
Sterile products and accessories are a crucial part of the ACCU-CHEK Spirit  
insulin pump system and therapy. These products are designed for safe and  
simple insulin pump therapy. Always follow the instructions given by your  
doctor or healthcare team for the use of all system components.  
Use your ACCU-CHEK Spirit insulin pump, sterile products, and accessories  
for the first time in the presence of your doctor or healthcare team.  
Sterile Products  
Each sterile product should be used only once. Reusing these products could  
lead to infection or could damage your pump. Do not use a sterile product if  
the package is damaged or opened.  
Sterility is guaranteed up to the indicated expiration date for all ACCU-CHEK  
sterile products that are in the original, unopened packaging.  
Your insulin pump was designed for use with U100 insulin.  
Only use sterile products and accessories that are designed  
!
L
for use with your insulin pump. The proper functionality  
of your insulin pump can only be guaranteed if you use  
ACCU-CHEK brand sterile products and ACCU-CHEK Spirit  
accessories. All ACCU-CHEK brand sterile products and  
accessories designed for the ACCU-CHEK Spirit insulin pump  
have been thoroughly tested and approved for use with  
your insulin pump. Other sterile products and accessories  
have not been tested for compatibility with your insulin  
pump and, if used, may therefore endanger your health  
or void your warranty.  
WARNING  
Chapter 1: Understanding Your New System  
ACCU-CHEK Spirit Insulin Pump  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Always have extra sterile products and accessories with you.  
This allows you to exchange components when needed.  
Materials designed for single use (such as cartridges and  
infusion sets) should not be reused due to the increased risk  
of infections, malfunctions, and/or incorrect insulin delivery.  
!
L
CAUTION  
See Disetronic Medical Systems’ sterile products and accessories catalogs and  
brochures for more information. Or, contact Pump Support at 1-800-688-4578.  
Your insulin pump and its sterile products and accessories  
include small component pieces that could pose a choking  
hazard to children. Keep sterile products and accessories  
out of the reach of children.  
!
L
WARNING  
Cartridges  
Your pump is designed to use ACCU-CHEK 3.15 ml plastic cartridges to hold  
insulin. Do not reuse a cartridge. Reusing sterile products could lead to infection  
or could damage your pump.  
ACCU-CHEK 3.15 ml plastic cartridges come with a protective cartridge cap,  
as well as a filling needle and plunger rod to aid in filling.  
cartridge  
plunger rod  
plunger  
cartridge protective cap  
filling needle with protective cap  
Chapter 1: Understanding Your New System  
18  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The cartridges are sterile products intended for single  
use only.  
!
L
WARNING  
Infusion Set  
An ACCU-CHEK infusion set connects your pump to your body to deliver insulin.  
The insulin travels from the cartridge in your insulin pump through the infusion  
set into your body. The cannula or needle of the infusion set is placed under  
the skin, usually in the abdominal area.  
Recommended  
Possible  
Chapter 1: Understanding Your New System  
ACCU-CHEK Spirit Insulin Pump  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Use only infusion sets that have a luer-lock connector. All current ACCU-CHEK  
infusion sets have luer-lock connectors and are the preferred choice for use  
with your ACCU-CHEK Spirit insulin pump. All ACCU-CHEK infusion sets are PVC  
free, are skin-friendly, and do not affect insulin. The diameter of the ACCU-CHEK  
infusion set tubing is small for fast priming and minimal waste of insulin.  
ACCU-CHEK infusion sets come in a wide range of lengths from 60 cm (24 inches)  
to 110 cm (43 inches). Choose your tubing length based on how you wear  
your pump. Please refer to the infusion set instructions for proper usage and  
priming volumes.  
Adapter  
Vents  
Seal inside the adapter  
Seal outside the adapter  
The adapter connects the cartridge to the infusion set. It has two seals to keep  
water from getting into the cartridge compartment. Two vents on the adapter  
allow air pressure to equalize. The adapter will not work properly if the vents are  
plugged, dirty, or if the seals are worn or missing.  
Change the adapter with at least every 10th cartridge change. Also change the  
adapter if the seals become worn, lost, or if the vents become plugged.  
Chapter 1: Understanding Your New System  
20  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Battery  
Your pump runs on one AA battery. Use AA alkaline batteries that have a minimum  
capacity of 2500 mAh. Do not use lithium, carbon zinc, or nickel cadmium (NiCd)  
batteries. If you prefer to use rechargeable batteries, use NiMH batteries that  
have a minimum capacity of 1500 mAh. If you use rechargeable batteries, always  
use a battery charger recommended by the battery manufacturer. Keep extra  
batteries on hand.  
Using batteries other than those recommended by Disetronic  
Medical Systems may significantly reduce battery usage  
time. Batteries other than those recommended could leak  
and corrode the battery contacts within your insulin pump.  
Therefore, using batteries not supplied or recommended  
by Disetronic Medical Systems, Inc. may void the warranty.  
!
L
CAUTION  
The operating temperature of the battery must be between 41°F and 104°F  
(+5°C and +40°C). The life of any battery is affected by how you use your  
insulin pump, your personal settings, delivery rates, temperature and other factors.  
With typical pump use [50 U/day using U100 insulin; operating temperature  
72°F 6°F (22°C 3°C)], pump battery life is about four weeks using alkaline  
batteries and one week using rechargeable batteries.  
Chapter 1: Understanding Your New System  
ACCU-CHEK Spirit Insulin Pump  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
You must set the battery type to match the kind of battery you are  
using. If you do not set the correct battery type on your pump, the  
alert A2: BATTERY LOW may not give you enough time to replace your  
battery. If you change from using regular batteries to rechargeable  
batteries, or from rechargeable batteries to regular batteries, you must  
change the battery type setting in the Standard setup menu (SETUP  
MENU STANDARD).  
Always make sure that you remove the battery if your  
insulin pump is not in use for a long time period in order  
to preserve the battery lifetime. To prevent water from  
entering the casing, change the battery only in a dry area.  
Make sure that the seal of the battery cover is not worn  
out or missing, and that the battery is properly inserted.  
!
L
CAUTION  
Battery Cover  
Seal  
The battery cover seals the battery compartment. Change the battery cover  
with at least every 4th battery change.  
• Remove the battery cover only in a dry area to prevent water from  
getting inside the pump.  
• Make sure that the seal is not worn or missing and that the battery  
is properly inserted into the pump.  
• Tighten or loosen the battery cover using only the ACCU-CHEK Spirit  
insulin pump battery key. Knives, screwdrivers, or other sharp objects  
may harm your pump.  
• Do not over-tighten the battery cover. The battery cover is correctly  
inserted and tightened when it is even with the pump casing.  
Chapter 1: Understanding Your New System  
22  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Battery Key  
The battery key tightens and loosens the battery cover. It can also loosen the  
luer-lock connection between the infusion set and adapter if it cannot be  
loosened by hand. However, never use the battery key to attach or to tighten  
an infusion set.  
The battery key has a notch that fits all ACCU-CHEK infusion set luer-lock  
connectors. The battery key should be replaced if it cracks, becomes worn,  
or if it is lost.  
Emergency Kit  
It is a good idea to carry supplies in case of an emergency so that you can  
exchange parts if you need to. An emergency kit might contain:  
• A new ACCU-CHEK infusion set  
• A new AA alkaline battery  
• A new ACCU-CHEK 3.15 ml plastic cartridge  
• A pen or syringe for alternate therapy  
• A vial of insulin  
• A filling aid  
• Treatment for low blood glucose levels  
• Blood glucose monitoring supplies (such as an ACCU-CHEK blood  
glucose monitor)  
• A disinfectant for skin  
• A new site dressing  
• The battery key  
This list is only an example of what an emergency kit might contain. Ask your  
doctor or healthcare team what supplies you should keep in your personal  
emergency kit.  
Chapter 1: Understanding Your New System  
ACCU-CHEK Spirit Insulin Pump  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
This page left blank intentionally.  
Chapter 1: Understanding Your New System  
24  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2: Getting Started  
Inserting (or Changing) the Battery  
Your pump turns on when you insert the battery. When you remove the battery,  
your pump still tracks the time and date for about 1 hour. The pump will always  
store your pump settings (such as the hourly basal rates, bolus increment and  
active user menu) and event memory (bolus and alarm history, history of daily  
insulin totals and Temporary Basal Rates) even if the battery power dies and no  
matter how long your pump is without power.  
See the section “Accessories and Disposables – Battery” (page 21) for information  
on recommended batteries.  
Keep the Time/Date Settings Correct  
!
L
Always check that the time and date of your insulin pump  
WARNING  
are set correctly. The time and date are displayed on the  
STOP screen. Incorrect programming of the time and date  
may cause incorrect insulin delivery. If you, your doctor,  
or healthcare team store and analyze your therapy data  
electronically and the time and date of the devices used  
are not set identically, the gathered data might not be  
meaningful. See the section “Setting Time and Date”  
(page 32) for more information.  
To prevent water from entering the casing, change the  
battery only in a dry environment. Make sure that the  
battery seal is not worn out or missing and that the  
battery is properly inserted. Tighten or loosen the battery  
cover with the ACCU-CHEK Spirit insulin pump battery key  
(knives, screwdrivers or other sharp objects may harm  
your insulin pump). Do not over-tighten as this may cause  
damage to the battery cover and to the casing. The battery  
cover is correctly inserted and tightened when the battery  
cover is level with the insulin pump casing.  
!
L
CAUTION  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Follow the steps below:  
1. Press d to move to the STOP YOUR ACCU-CHEK screen.  
Press f to select and put the pump in STOP  
mode. Your pump returns to the STOP screen  
and insulin delivery stops.  
2. Remove or disconnect your infusion tube set from the headset so that insulin  
cannot accidentally be delivered.  
3. Remove the battery cover.  
Use the battery key to turn the battery cover as  
indicated by the arrow on the battery cover. Make  
sure that the opening of the battery compartment  
and the seal are clean and undamaged. Remove  
the old battery if necessary.  
4. Put the battery cover on the positive (+) end of the battery.  
Hold the cover on the positive (+) end of the battery,  
and insert the battery negative (-) end first into the  
battery compartment.  
Chapter 2: Getting Started  
26  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Use the cover to gently push the battery all the way in.  
Turn the battery cover clockwise to tighten it.  
The battery cover is correctly tightened when  
the outside of it is even with the casing. Do not  
over-tighten the battery cover.  
The pump begins its start-up process.  
Note:  
If your pump was in RUN mode before you removed the battery, an  
error E8: POWER INTERRUPT will occur when a new battery is inserted.  
Press f twice to confirm and turn off the error. Check the time and  
date in the STOP screen and correct them, if necessary. See the section  
“Setting Time and Date” (page 32) for instructions. Put your pump into  
RUN mode if necessary.  
Using AA batteries other than those recommended by  
Disetronic Medical Systems may significantly reduce the  
!
L
battery usage time. Batteries other than those recommended  
may leak and corrode the battery contacts within your  
insulin pump. Therefore, using batteries not recommended  
by Disetronic Medical Systems may void the warranty. Do  
not insert old or used batteries into your insulin pump as  
this may cause a failure at the insulin pump start-up.  
CAUTION  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
System Check Start-Up Process  
Your pump performs internal tests at start-up when you insert a battery and when  
you change the cartridge. When you insert a battery the whole start-up process  
is performed. When only the cartridge is changed, the start-up process begins  
at the SELF-TEST screen.  
Do not interrupt the start-up process by pressing keys  
or by handling the insulin pump. Interruption of the  
start-up process may lead to malfunctions of your  
insulin pump.  
!
L
WARNING  
If your insulin pump does not beep and vibrate as  
described, or if individual numbers, letters, symbols  
or lines are shown incompletely or not at all, contact  
Pump Support at 1-800-688-4578.  
Note:  
While you are using the pump, its safety system monitors it continuously.  
If the safety system detects a change from the normal state, an alert  
(warning instruction) or an error (error message) occurs.  
1. Put the pump in STOP mode.  
2. The ACCU-CHEK logo appears.  
Chapter 2: Getting Started  
28  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. SELF-TEST appears.  
4. Your pump beeps and BEEP TEST appears.  
Listen for the beeps.  
If you cannot hear the beeps from your insulin pump, you  
!
may not be notified of changes with your insulin pump in  
a suitable amount of time. If you cannot hear the beeps,  
contact Pump Support at 1-800-688-4578.  
L
CAUTION  
5. Your pump vibrates and VIBRATION TEST appears.  
Check for vibrations.  
If you cannot feel the vibration alert, you may not be  
notified of any changes of your insulin pump in a suitable  
CAUTION amount of time. If you cannot feel the vibration alert,  
contact Pump Support at 1-800-688-4578.  
!
L
6. The display goes blank. Check that the display is empty. (The backlight turns  
on during this step.)  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. The display shows a pattern. Check that the pattern is even.  
8. The display turns black. Check that the display is completely black.  
9. You will also see an INFORMATION screen.  
10. If your pump detects an alert or error during the start-up process, it  
signals the error or alert at this point. See the section “Alerts and Errors”  
(page 147) for more information.  
11. A melody signals the end of the start-up process, and the pump shows  
the STOP screen. The STOP-Warning occurs every minute by giving a  
long beep and a vibration to remind you that no insulin is being delivered  
in STOP mode. Press and hold a or s until you hear a melody to turn  
the STOP-Warning off.  
Chapter 2: Getting Started  
30  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
If the voltage of the battery is too low, the start-up procedure  
cannot finish.  
Your pump beeps five tones and keeps restarting the start-up  
procedure until you remove the battery or until the battery is dead.  
If you discover incomplete characters, numbers, or symbols  
on the display, put your insulin pump into STOP mode and  
!
L
CAUTION  
perform a system check again by simply removing the battery  
for a few seconds. When you reinsert the battery the system  
check will occur. If the incomplete characters, numbers, or  
symbols remain, your insulin pump may need to be returned  
since you may not receive the correct information for  
operating your pump. Put your insulin pump into STOP mode  
and contact Pump Support at 1-800-688-4578 immediately.  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Setting Time and Date  
Be careful to set the time and date correctly, because the basal rates and all  
history information are stored on the basis of time and date set in the pump.  
When you travel across time zones, make sure that the time and date are  
set correctly.  
The time format can be either American or European/Military Time.  
American  
12-hour clock  
month/day/year  
European/Military Time  
24-hour clock  
day/month/year  
Incorrect setting of time and date can lead to incorrect  
insulin delivery. Make sure that the time and date of your  
insulin pump is programmed correctly to ensure the  
correct insulin delivery and data memory. Do not make  
therapy decisions based upon a single result in the  
insulin pump’s memory.  
!
L
WARNING  
If you, your doctor, or healthcare team reviews your therapy  
data electronically, it is essential that the time and date  
of your insulin pump, your blood glucose measurement  
system (such as an ACCU-CHEK blood glucose meter),  
the PC, and other devices used are set identically. If they  
are not, the gathered data might not be meaningful. Check  
these devices regularly to ensure that the time and date  
settings are identical.  
Chapter 2: Getting Started  
32  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RUN  
screen  
set the time  
and date  
set the  
hour  
set the  
minute  
set the  
year  
set the  
month  
set the  
day  
Follow the steps below to set the Time and Date:  
1. Press d to move to the SET TIME AND DATE screen.  
Press f to select.  
2. The SET HOUR screen displays.  
Press a or s to set the hour.  
3. Press d to move to the SET MINUTE screen.  
Press a or s to set the minutes.  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Press d to move to the SET YEAR screen.  
Press a or s to set the year.  
5. Press d to move to the SET MONTH screen.  
Press a or s to set the month.  
6. Press d to move to the SET DAY screen.  
Press a or s to set the day.  
7. Press f to save and exit.  
Chapter 2: Getting Started  
34  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Your Personal Settings  
Your new ACCU-CHEK Spirit insulin pump must be  
!
programmed with your personal settings prior to starting  
insulin pump therapy. Work with your doctor or healthcare  
team to understand how to use the pump.  
L
WARNING  
Always check your personal settings before using your  
new pump in order to avoid incorrect insulin delivery.  
Record your personal settings from your current insulin  
pump and ensure that they are correctly programmed  
into a new ACCU-CHEK Spirit insulin pump. Incorrect  
programming of your insulin pump may cause incorrect  
insulin delivery.  
Do not operate your insulin pump without knowing your  
personal settings. If you are not sure about your personal  
settings or you are less experienced, have your doctor or  
healthcare team check them. Work with your doctor or  
healthcare team to determine the timing as well as the  
amount and type of bolus you need to deliver. Make sure  
you know your personal carbohydrate to insulin ratio and  
your correction bolus ratio.  
Change your personal settings only after consulting with your  
doctor or healthcare team and always follow their instructions.  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Your Basal Rate Profile  
Your pump delivers insulin every 3 minutes, in 20 equal doses each hour,  
24 hours a day. This flow of insulin, measured in international units per hour  
(U/h), is called the basal rate and is calculated to meet your basic insulin needs.  
An ACCU-CHEK Spirit insulin pump basal rate profile consists of up to 24  
different hourly basal rates. Each hourly rate may be changed independently.  
The total of all 24 hourly basal rates in one basal rate profile is called the  
(daily) basal rate total.  
Your basal rate profile should be reviewed:  
• after any programming changes are made to your pump,  
• after a new battery is inserted, and  
• after confirming an error E7: ELECTRONIC ERROR.  
Chapter 2: Getting Started  
36  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Programming Your Basal Rate Profile  
Note:  
Insulin delivery does not stop or pause during basal rate programming.  
The insulin amounts and other values used in this user guide are  
examples only. Your personal values may differ.  
RUN  
screen  
program  
profile 1  
daily basal  
rate total  
1st hour  
2nd hour  
24th hour  
Note:  
For your safety and convenience during programming, your insulin  
pump returns itself to the RUN or STOP screen if no key is pressed  
within 20 seconds. The changes you made before your insulin pump  
“times out” will not be saved.  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Follow the steps below:  
1. Press d to move to the PROGRAM BASAL RATE PROFILE 1 screen.  
2. Press f to select.  
The daily BASAL RATE TOTAL is displayed.  
3. Press d to move to the first hour.  
The first hour, which always begins at midnight,  
00:00 – 01:00 (12:00 AM – 01:00 AM if the  
American time format is used), is shown.  
4. Press a or s or scroll to set the hourly basal rate given by your doctor  
or healthcare team.  
5. Press d to move to the next hour.  
6. Continue using d and the a and s keys to set the hourly basal rates  
for the remaining hours. This lets you program your individual basal rate  
profile hour-by-hour. Continue until all 24 hours are programmed.  
Chapter 2: Getting Started  
38  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Press f to confirm.  
The new daily BASAL RATE TOTAL is displayed.  
Check the new daily BASAL RATE TOTAL.  
8. Press f to save and exit.  
If the current active basal rate profile is the same as the new programmed  
basal rate, it is active immediately.  
Note:  
At any screen you have three options to exit:  
If you want to save the changes  
• Press f. Check the new daily basal rate total.  
• Press f. Your pump returns to the RUN screen.  
If you want to undo the changes  
• Wait for your ACCU-CHEK Spirit insulin pump to return to the RUN  
screen (timeout), or  
• Press d+a simultaneously (exit feature) to exit the function screen.  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Copying an Hourly Basal Rate  
If your basal rate is the same for multiple hours in a row, you can save time  
programming by copying an hourly basal rate to one or more hours.  
Follow the steps below:  
1. Press d to move to the PROGRAM BASAL RATE PROFILE 1 screen.  
2. Press f to select.  
The daily BASAL RATE TOTAL is displayed.  
3. Press d to move to the hourly basal rate you want to copy.  
4. Press a+s simultaneously.  
A special beep confirms that you are pressing  
the keys correctly. The same hourly basal rate  
is copied to the next hour.  
5. If you want to copy the rate to multiple hours, press a+s repeatedly or  
press and hold a+s (scroll).  
Chapter 2: Getting Started  
40  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Press f to confirm.  
Check the new daily BASAL RATE TOTAL.  
7. Press f to save and exit.  
If the current active basal rate profile is the same as the new programmed  
basal rate, it is active immediately.  
Basal Rate and Bolus  
During normal use, your pump is in RUN mode and insulin is continuously  
delivered (basal rate).  
A bolus can be programmed when it is needed. A bolus is the amount of insulin  
delivered (in addition to the basal rate) to cover the intake of food and correct  
high blood glucose levels. The bolus amount is determined by your doctor or  
healthcare team’s guidelines, your blood glucose level, your food intake, and  
your activity level as well as stress and illness. See the section “Understanding  
Boluses” (page 75) for more information.  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Preparing the Cartridge  
Your insulin pump is designed to use an ACCU-CHEK  
!
3.15 ml plastic cartridge. The cartridge is a sterile product  
intended for single use only. Sterility is guaranteed for  
unopened packaging up to the indicated expiration date.  
L
WARNING  
Do not use sterile products if the package is damaged.  
Do not reuse single use materials. Reusing single use  
materials might lead to a malfunction of your insulin  
pump and to incorrect insulin delivery and/or an infection.  
Always handle the items with clean hands.  
Use specified cartridges only. Using other cartridges  
can endanger your health and may void the warranty.  
Chapter 2: Getting Started  
42  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Filling the Cartridge  
Note:  
A filling aid, an optional accessory by Disetronic Medical Systems, helps  
you fill an empty ACCU-CHEK 3.15 ml plastic cartridge from your vial of  
insulin. Read the instructions that are included with the filling aid to learn  
how to use it.  
To fill a cartridge, have the following materials ready:  
A new 3.15 ml plastic cartridge with protective cap, plunger rod,  
filling needle, and  
Insulin at room temperature to minimize formation of air bubbles  
Follow the steps below:  
1. Wash your hands.  
2. Set the insulin vial on a hard, flat surface, such as a table.  
Clean the rubber membrane of the vial using an  
antiseptic wipe.  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Remove the cartridge, the cartridge protective cap, and the filling needle  
with its protective cap from their packaging.  
4. Place the filling needle with protective cap onto the cartridge tip and make  
sure the connection is tight.  
Move the plunger rod back and forth twice in the  
cartridge to distribute the lubricant.  
Pull the plunger back to fill the cartridge with air.  
Plunger  
Plunger rod  
5. Remove the protective cap from the needle. Do not touch anything with  
the needle.  
Push the tip of the filling needle into the center  
of the insulin vial’s rubber membrane.  
Push the plunger rod down so that all air from  
the cartridge enters the insulin vial.  
Chapter 2: Getting Started  
44  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Hold steady pressure against the plunger rod with your thumb, and at the  
same time turn the insulin vial over so that the filling needle and cartridge  
are pointing upwards into the insulin vial.  
Make sure that the tip of the filling needle stays in the insulin.  
Slowly release the pressure against the plunger rod and  
let the insulin flow into the cartridge.  
Do not pull or push the plunger rod while it is moving on  
its own (this could lead to air bubbles in the insulin).  
Slowly pull the plunger rod straight down to fill the cartridge.  
Remove any air bubbles by tapping on the cartridge. Push the  
air bubbles back into the insulin vial using the plunger rod.  
The cartridge is full when there are no air bubbles in it and  
the plunger is at the bottom of the cartridge.  
7. Remove the filling needle from the insulin vial.  
8. Place the protective cap on the filling needle.  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Remove the plunger rod by turning it counter-clockwise  
out of the plunger.  
Remove the filling needle with its protective cap from the  
cartridge by turning counter-clockwise. Dispose of properly.  
Push the cartridge protective cap tightly onto the cartridge  
tip until it clicks.  
The cartridge is now ready to use.  
Note:  
Do not pull or push the plunger rod during removal.  
Air bubbles in the cartridge and infusion set may cause  
the infusion of air instead of insulin. In this case, your body  
does not receive the required amount of insulin and an  
error E4: OCCLUSION may be displayed. Remove bubbles  
while filling the cartridge and priming the infusion set and  
without having the infusion set connected to your body.  
!
L
CAUTION  
Cold insulin may release air when warmed. Use only insulin  
at room temperature when filling the cartridge and priming  
the infusion set. Inspect the cartridge and the infusion set for  
air bubbles at least every three hours during the day and  
before you go to sleep. Remove any air bubbles and change  
system components if needed.  
Chapter 2: Getting Started  
46  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Connecting the Cartridge, Adapter, and Infusion Set  
Note:  
Before you change the cartridge, adapter, and infusion set, check  
that your pump is in good, working order. See the section “System  
Inspection Checklist” (page 137).  
Insulin can flow freely from the cartridge or infusion set if the cartridge  
plunger and the piston rod are not properly connected and the pump  
is positioned at a higher level than the infusion site. Prevent free flow  
of insulin by correctly inserting the cartridge: move the piston rod to  
the proper position and twist the adapter until the cartridge plunger  
sits flush with the end plate of the piston rod.  
Have the following materials ready:  
A filled ACCU-CHEK 3.15 ml plastic cartridge with a luer-lock connection  
An ACCU-CHEK Spirit insulin pump adapter  
A new ACCU-CHEK infusion set  
The cartridge and infusion set are sterile products intended  
for single use only. Sterility is guaranteed for unopened  
packaging up to the indicated expiration date. Do not use  
sterile products if the packaging is damaged.  
!
L
WARNING  
Do not reuse single use materials because this might lead  
to a malfunction of your insulin pump, to incorrect insulin  
delivery, and/or an infection. Always handle the items with  
clean hands. Avoid any contact of the infusion set and the  
connecting parts of your insulin pump with antiseptics,  
antibiotic creams, soaps, perfumes, deodorants, body  
lotions, or any other cosmetics. They may contaminate  
these parts.  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Follow the instructions below:  
Remove the protective cap from the cartridge. Do not touch the cartridge tip.  
Push the adapter all the way onto the cartridge tip to  
b
the stop (a).  
a
Luer-lock connector  
Hold the adapter and twist the infusion set luer-lock  
connection by hand, clockwise into the adapter (b).  
Carefully prepare a new ACCU-CHEK infusion set for use,  
priming and insertion, following its instructions for use.  
Do not over-tighten.  
To avoid leakage, properly tighten the infusion set to the  
adapter. Turn the infusion set luer-lock connector until it  
stops. Do not turn it any further and do not use any auxiliary  
tools as this may crack the infusion set luer-lock connector  
and result in leakage.  
!
L
CAUTION  
Note:  
Your pump cannot detect infusion set leakage. You must check all parts of  
the infusion set at least every three hours during the day and before you go  
to sleep. If you detect any loss of insulin and all parts are correctly tightened,  
immediately replace the leaking component. Because insulin delivery  
has been interrupted, check your blood glucose level and take appropriate  
actions according to your doctor’s or healthcare team’s instructions.  
Chapter 2: Getting Started  
48  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Inserting the Cartridge  
Do not perform the CHANGE THE CARTRIDGE function if the  
!
cartridge compartment is not completely dry. During the  
rewinding of the piston rod, insulin or water might enter your  
insulin pump and could lead to malfunction. If necessary,  
use a soft cloth to dry the compartment.  
L
CAUTION  
3 sec.  
STOP  
screen  
change the  
cartridge  
remove the  
cartridge  
returning  
piston rod  
cartridge volume  
315 units  
start-up  
procedure  
prime the  
infusion set  
Have the following materials ready:  
Your insulin pump  
A filled ACCU-CHEK 3.15 ml plastic cartridge with a connected  
ACCU-CHEK Spirit insulin pump adapter  
A new ACCU-CHEK infusion set  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Follow the steps below:  
1. Press d to move to the STOP YOUR ACCU-CHEK screen.  
2. Press f to select and put the pump in STOP mode.  
Your pump displays the STOP screen, and  
insulin delivery stops.  
3. Press d to move to the CHANGE THE CARTRIDGE screen.  
4. Press f to select.  
Remove the current cartridge, if there is one.  
5. Press and hold f (for 3 seconds) until you hear a melody.  
The piston rod begins its return. Do not touch or  
disturb the piston rod while it is rewinding. Doing  
so may potentially damage your pump.  
When the piston rod returns completely, the  
CARTRIDGE VOLUME screen displays the  
maximum cartridge volume, 315U.  
Chapter 2: Getting Started  
50  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Hold your pump upright. Hold the new filled cartridge so that the connected  
adapter and infusion tube set point upwards, parallel, and close to the  
cartridge compartment.  
Make sure that the bottom edge of the colored  
part of the adapter is level with the top of the  
cartridge compartment.  
Base plate of  
piston rod  
Cartridge plunger  
7. Press a or s to move the plunger rod forward until the end plate of  
the piston rod is level with the bottom of the cartridge plunger.  
The cartridge volume decreases accordingly  
on the display.  
8. Position your insulin pump upright (with the adapter pointing upwards).  
Insert the cartridge into the cartridge compartment.  
9. Twist (do not push) the adapter clockwise until it sits flush with the cartridge  
compartment.  
It is not necessary to apply pressure. Twisting will  
properly position the cartridge. Do not over-tighten  
the adapter or it will be difficult to remove the  
cartridge, adapter and cartridge plunger and may  
potentially cause damage to your pump. The  
cartridge is correctly inserted when the end plate  
of the piston rod sits flush against the cartridge  
plunger. This flush fit is an extra safety measure  
to prevent the free flow of insulin and to minimize  
occlusions (blockages).  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
If the end plate of the piston rod does not sit flush against the cartridge plunger,  
remove the cartridge from the cartridge compartment. Hold the cartridge  
and adapter next to the cartridge compartment. Press a or s to move  
the piston rod forward until its base plate and the cartridge plunger are  
level. The displayed cartridge volume decreases accordingly. Re-insert the  
cartridge into the cartridge compartment.  
– If needed, repeat this step until the cartridge is correctly inserted (the  
end plate of the piston rod sits flush against the cartridge plunger).  
10. Press f after correctly inserting the cartridge.  
Your pump performs a self test. See the section  
“System Check Start-Up Process” (page 28).  
11. Check that the infusion set luer-lock connector is still correctly connected  
to the adapter. Tighten the infusion set by hand clockwise until it sits tightly  
in the adapter.  
12. When the system check start-up process is finished, the PRIME THE INFUSION  
SET screen is displayed.  
See the section “Priming the Infusion Set” (page 53) for more information.  
Start from the point in those instructions where the pump screen matches  
the screen above.  
Chapter 2: Getting Started  
52  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Priming the Infusion Set  
Never prime an infusion set that is connected to your body.  
You risk uncontrolled insulin delivery into your body. Always  
follow the instructions for the infusion set you are using.  
!
L
WARNING  
The ACCU-CHEK Spirit insulin pump stops priming after 25 units of insulin. If no  
insulin has emerged from the tube set needle after priming, repeat the priming  
function. When insulin comes from the tip of the needle, press any pump key  
to stop priming. The amount of insulin used in priming is not added to the pump’s  
history of daily insulin totals.  
During priming your pump should be upright, with the adapter pointing upwards  
to help let air bubbles out of the cartridge and the infusion set.  
Air bubbles in the cartridge and infusion set may cause the  
infusion of air instead of insulin. If this happens, your body  
!
L
will not receive the required amount of insulin. Inspect the  
cartridge and the infusion set for air bubbles at least every  
three hours during the day and before you go to sleep.  
Remove bubbles while filling the cartridge and priming the  
infusion set and without having the infusion set connected  
to your body.  
WARNING  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
start-up  
procedure  
3 sec.  
stop  
priming  
STOP  
screen  
prime the  
infusion set  
start  
priming  
Follow the steps below:  
1. Press d to move to the STOP YOUR ACCU-CHEK screen.  
2. Press f to select and put the pump in STOP mode.  
Your pump displays the STOP screen, and insulin  
delivery stops.  
3. Prepare the tubing or cannula/needle for priming per infusion set instructions  
for use.  
4. Press d to move to the PRIME THE INFUSION SET screen.  
Chapter 2: Getting Started  
54  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Press f to select. The priming amount of 25 units of insulin is shown on  
the display.  
6. When START PRIMING is displayed, press and hold f until you hear a  
melody and the piston rod moves forward.  
The priming process begins and a count-up is  
shown on the display. Your pump stops priming  
after 25 units.  
When the priming process is finished, your pump  
returns to STOP.  
The infusion tubing set is properly primed when  
there are no air bubbles visible in the tubing and a drop of insulin has come  
out of the connector or cannula.  
Stopping the Priming Process  
You can stop the priming process at any time by  
pressing any key. Your pump displays the STOP  
PRIMING screen.  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Preparing the Infusion Site and Inserting the Infusion Set  
Always follow the instructions given to you by your doctor or healthcare team  
and the instructions for the infusion set you are using.  
Change your infusion set as frequently as recommended by the manufacturer.  
Infusion Site Selection  
Your doctor or healthcare team will help you select your infusion site areas  
according to a planned rotation pattern. Avoid your waistline, bones, scar tissue,  
recent infusion sites, bruises, and sores. Choose an infusion site at least one inch  
(2.5 cm) away from the navel and previous infusion sites.  
Infusion Site Preparation  
It is important to prepare the infusion site properly to reduce the risks of infection.  
Your doctor or healthcare team will give you infusion site preparation guidelines.  
Have the following materials ready:  
Your ACCU-CHEK Spirit insulin pump  
A disinfectant for your skin  
A new ACCU-CHEK infusion set  
Follow the steps below:  
1. Make sure your pump is correctly set up (including cartridge, adapter and  
infusion set) and is programmed with your personal settings.  
Chapter 2: Getting Started  
56  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Wash your hands thoroughly.  
3. Disinfect the infusion site and let it dry completely.  
4. Make sure your pump is in STOP mode.  
5. Insert your ACCU-CHEK infusion set according to the instructions for its use.  
Chapter 2: Getting Started  
ACCU-CHEK Spirit Insulin Pump  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
If you use a soft cannula-type infusion set, you must proceed with a bolus  
to fill the air space in the cannula after removing the introducer needle  
according to the instructions for the infusion set you are using. Failure to  
do so could result in missed insulin dosing. See the section “Programming  
a Standard Bolus” (page 77) for more information.  
Inspect your infusion site at least once or twice each day for  
irritation or infection. Signs of infection may include, but  
are not limited to, pain, lumps, redness, heat, or discharge.  
!
L
CAUTION  
If you see redness or swelling, change your infusion headset  
and infusion site immediately and contact your doctor or  
healthcare team.  
Change your infusion headset according to the instructions  
for the infusion set you are using and your doctor or  
healthcare team’s recommendations.  
If you have read and followed all instructions, cautions, warnings, and notes  
found in this chapter, your pump should now be ready to use. However, please  
read the complete manual before starting.  
Chapter 2: Getting Started  
58  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3: Using Your Pump  
Before Using  
Training and using your insulin pump requires the support of  
an experienced doctor or healthcare team. Regular visits  
to your doctor or healthcare team are absolutely essential  
during insulin pump therapy.  
!
L
CAUTION  
Always follow the instructions given by your doctor or  
healthcare team.  
Successful insulin pump therapy requires you to check your blood glucose  
levels often. It is a good idea to check your blood glucose (using an ACCU-CHEK  
blood glucose monitor) at least four times a day, or as often as you are  
directed by your doctor or healthcare team.  
Wearing Your Pump  
When wearing your insulin pump make sure to prevent  
any contact with objects that may damage or accidentally  
press the keys of your insulin pump (e.g., keychains, keys,  
pocket knives, and coins).  
!
L
CAUTION  
Attach your pump safely to your body or clothes to prevent damage. Special  
carrying systems let you carry your pump on or under your clothes. All specified  
accessories have been tested and approved for use with your ACCU-CHEK Spirit  
insulin pump. The proper functioning of your pump can only be guaranteed if you  
use it with ACCU-CHEK Spirit insulin pump accessories.  
See Disetronic Medical Systems sterile products and accessories catalogs and  
brochures for more information. Or, contact Pump Support at 1-800-688-4578.  
Chapter 3: Using Your Pump  
ACCU-CHEK Spirit Insulin Pump  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Wear your pump under your clothes or directly on your body in cold or rainy  
weather. If you have concerns about the weather affecting your pump,  
check the operating conditions of your pump and contact Pump Support at  
1-800-688-4578. See the section “Appendix A: Technical Data” (page 165) for  
more information on operating conditions.  
Do not place your insulin pump in direct sunlight. Overheating  
of the insulin and your insulin pump must be avoided. Protect  
your insulin pump from direct exposure to cold wind and  
temperatures over 104°F (40°C) and below 41°F (5°C). These  
conditions may damage the insulin and the electronics of  
your insulin pump and may cause a malfunction of the battery.  
!
L
CAUTION  
See the instructions for the insulin you are using to determine  
the acceptable temperature range.  
Chapter 3: Using Your Pump  
60  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
When Not to Wear Your Pump  
Electromagnetic Fields and Hazardous Areas  
Do not use your pump around electromagnetic fields such as radar or antenna  
installations, high-voltage sources, X-Ray sources, MRI, CAT scan, and all other  
sources of electrical current as they may cause your pump not to work. Insulin  
delivery may stop and an error E7: ELECTRONIC ERROR may occur.  
Always stop and remove your pump before you enter these areas.  
See “Appendix A: Technical Data” (page 165) for more information about  
electromagnetic fields.  
Your insulin pump has not been tested with cardiac  
pacemakers. Do not use the ACCU-CHEK Spirit insulin  
pump if you have a cardiac pacemaker.  
!
L
WARNING  
Your pump has been tested and complies with the regulations regarding  
unintentional electromagnetic interference. Security systems in airports as  
well as anti-theft monitoring devices, such as those in department stores,  
should not affect the functioning of your pump.  
Because there are so many devices, such as cellular phones, that give out  
electromagnetic radiation, the influence of some such devices on your pump  
cannot be ruled out. It is a good idea to keep your pump at least 4 inches  
(10 cm) away from such devices while the other devices are turned on.  
Do not use your pump in hyperbaric chambers and in hazardous areas of any  
kind, such as areas where explosive or flammable gases or vapors could exist.  
This could affect insulin delivery and/or lead to harm.  
Your pump is designed to work in normal barometric conditions (from 700  
to 1,060 mbar).  
Chapter 3: Using Your Pump  
ACCU-CHEK Spirit Insulin Pump  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Do not exceed 10,000 feet (3,000 meters) above sea level. Your pump has  
not been tested for use above this level.  
Always stop and remove your pump before you enter these areas. If you  
have questions, contact Pump Support at 1-800-688-4578.  
Sports and Exercise  
Exercise is an important part of diabetes management. Disetronic Medical  
Systems designed the ACCU-CHEK Spirit insulin pump to let you safely perform  
many sporting activities while you wear your pump. It is important that you  
protect your pump before you start exercise or sports activities. Do not wear  
your pump during sports that involve body contact such as boxing, football,  
or hockey since rough contact could damage your pump.  
Consult our sterile products brochures and accessories catalog for more  
information regarding carrying systems or contact Pump Support at  
1-800-688-4578 for more information.  
Traveling  
Ask your doctor or healthcare team about any special actions you need to  
take before you travel. Take extra blood-glucose testing and pump supplies  
when you travel and find out where to get supplies while you are traveling.  
When you travel across time zones, make sure that the time and date on  
your pump are set correctly to the local time.  
The ACCU-CHEK insulin pump has not been tested at altitudes in excess of  
10,000 feet (3,000 meters) above sea level. If your travels take you to an area  
in excess of 10,000 feet (3,000 meters) above sea level, stop and remove your  
pump. This does not apply to use in airplanes with pressurized cabins.  
If you have questions, contact Pump Support at 1-800-688-4578.  
Chapter 3: Using Your Pump  
62  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Your Insulin Pump and Water  
Avoid deliberate contact with water. Check daily that your  
!
insulin pump is not chipped, cracked, or damaged in any  
way and that the battery cover and the adapter are correctly  
closed. In the presence of chips and cracks, water, dust,  
insulin, or other foreign substances may enter your insulin  
pump and lead to malfunction. Prior to contact with water,  
disconnect and take off your insulin pump.  
L
CAUTION  
Disconnect and remove your pump before entering a bathtub, Jacuzzi or  
whirlpool, shower, or swimming pool. Avoid exposing your pump to high  
humidity, such as in a sauna, as this could damage your pump. To make it  
easy to remove your pump, use disconnectable ACCU-CHEK infusion sets.  
Consult your doctor or healthcare team to find out how long your insulin pump  
therapy can safely be interrupted.  
Accidental Contact with Water  
Your insulin pump is protected against accidental short-term water contact.  
Examples of accidental short-term water contact include:  
• rain, snow;  
• water that splashes while you are biking, jogging, or hiking;  
• accidental brief immersion into the sink or bathtub.  
Chapter 3: Using Your Pump  
ACCU-CHEK Spirit Insulin Pump  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
What to Do After Any Water Contact  
Put your pump into STOP mode. Disconnect the pump from your body before  
inspecting it. Use a soft cloth to dry the outside of the pump. Check the battery  
compartment and the cartridge compartment for water. If the battery compartment  
or cartridge compartment are wet, turn your pump upside down to let the water  
run out and let it air dry. Do not use warm air such as a hair dryer to dry the pump  
as this could damage the pump’s electronics. Do not re-insert the battery or the  
cartridge until the compartments are completely dry.  
See the section “Cleaning Your ACCU-CHEK Spirit Insulin Pump” (page 141)  
for more information.  
Other Liquids  
You do not have to worry about contact between your pump and:  
• sweat  
• saliva  
You must check your pump immediately after contact with other liquids or  
chemicals such as:  
• cleaning solutions  
• alcohol  
• beverages  
See the section “Cleaning Your ACCU-CHEK Spirit Insulin Pump” (page 141)  
for more information.  
Avoid any contact of your insulin pump, the infusion set,  
and the connecting parts of your insulin pump with health  
and beauty items (for example, antiseptics, antibiotic  
creams, soaps, perfumes, deodorants, body lotions, or any  
other cosmetics). These items may discolor your pump or  
fog your display.  
!
L
CAUTION  
Chapter 3: Using Your Pump  
64  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Starting Insulin Delivery  
STOP  
screen  
start your  
ACCU-CHEK  
RUN  
screen  
Follow the steps below:  
1. Start from the STOP screen.  
2. Press d to move to the START YOUR ACCU-CHEK screen.  
3. Press f to select. Your pump shows the RUN screen.  
Insulin delivery begins within 3 minutes.  
Note:  
Check the amount of insulin remaining in the cartridge at least once a  
day. If you press the f key from the RUN or STOP screen, your pump  
displays the remaining cartridge content in units. Before going to sleep,  
make sure that:  
• the cartridge contains enough insulin to last through the night, and  
• the time and date are set correctly.  
Chapter 3: Using Your Pump  
ACCU-CHEK Spirit Insulin Pump  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Stopping Insulin Delivery  
When your insulin pump is in STOP mode, it does not deliver  
any insulin. You will need to put your insulin pump into RUN  
mode to continue insulin delivery.  
!
L
CAUTION  
There are functions and actions, for your safety, that you can only perform  
when the pump is in STOP mode. These functions and actions include:  
• changing the cartridge  
• connecting and disconnecting the adapter and/or infusion set, and  
• priming the infusion set  
Follow the steps below:  
1. Press d to move to the STOP YOUR ACCU-CHEK screen.  
2. Press f to select.  
Your pump returns to the STOP screen and  
insulin delivery stops.  
3. Press f (QUICK INFO screen) to review activated functions and features  
(like the beep and vibration alarm signals) indicated by icons and messages  
on the screen.  
Chapter 3: Using Your Pump  
66  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
When an Extended Bolus, a MultiWave Bolus, and/or a Temporary Basal  
Rate (TBR) is/are active and you put your pump into STOP mode, an alert  
A6: TBR CANCELLED and/or an alert A8: BOLUS CANCELLED occur(s).  
Press f twice for each appearing alert to confirm and turn off. See the  
sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8: BOLUS  
CANCELLED” (page 153) for more information.  
STOP-Warning  
The STOP-Warning is turned on by the pump each time your pump switches from  
RUN mode to STOP mode, and when a battery is inserted. The STOP-Warning  
is a long beep and a vibration occurring every minute. It warns you that insulin  
delivery has stopped. If you are aware of the situation and don’t wish to be  
reminded, you can turn off the STOP-Warning.  
To Turn Off the STOP-Warning  
Press either a or s for 3 seconds to turn off the STOP-Warning. A melody  
plays indicating the STOP-Warning is turned off.  
Note:  
The volume of the STOP-Warning cannot be changed. It is independent of  
the programmed beep volume.  
Chapter 3: Using Your Pump  
ACCU-CHEK Spirit Insulin Pump  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Interruption of Insulin Pump Therapy  
Consult with your doctor or healthcare team about when and for how long your  
insulin pump therapy can be interrupted.  
Measure your blood glucose level regularly during any insulin interruption. Use  
a syringe or insulin pen to inject insulin according to the instructions of your  
doctor or healthcare team.  
Since only rapid-acting insulin is used for insulin pump  
therapy, there is only a small insulin reserve in the  
body. If insulin is interrupted for any reason, such as:  
!
L
WARNING  
• if you stop the pump  
• there is a technical problem with the pump  
• there is leakage in the cartridge  
• there is occlusion in the infusion tube set or in the  
infusion headset  
• the cannula has slipped out of the infusion site  
you must be ready to replace the missing insulin  
immediately.  
Always carry spare sterile products and accessories  
(infusion set, insulin cartridge, batteries), an insulin  
pen/syringe, and insulin.  
Without insulin, diabetic ketoacidosis may develop  
and may require in-patient hospital treatment.  
Chapter 3: Using Your Pump  
68  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Short Interruption  
1. Put your pump in STOP mode.  
2. Remove or disconnect the infusion tube set from the infusion headset. When  
using a disconnectable infusion set, disconnect it at your infusion site and  
use the protective covers that came with the infusion set.  
Resuming Insulin Pump Therapy  
1. Attach a new ACCU-CHEK infusion set and prime it before inserting it.  
2. Follow the instructions to prepare the infusion set you are using, and put your  
pump into RUN mode.  
Measure your blood glucose level within 2 hours of restarting pump therapy to  
check the function of the system and to ensure insulin delivery.  
Interruption of Use of Your Insulin Pump for a Long Period of Time  
There may be times when you may need to interrupt your pump therapy for more  
than one day. Follow the steps below.  
Contact your doctor or healthcare team for an alternate  
therapy plan when you interrupt the insulin pump therapy  
for a long period of time.  
!
L
CAUTION  
1. Put your pump into STOP mode.  
2. Remove the cartridge, adapter, and infusion set.  
3. Reinsert the adapter.  
4. Remove the battery and reinsert the battery cover.  
5. Store your pump properly. See the section “Storing Your ACCU-CHEK Spirit  
Insulin Pump” (page 140) for more information.  
Chapter 3: Using Your Pump  
ACCU-CHEK Spirit Insulin Pump  
69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Changing the Infusion Set  
When you change the infusion set, keep the following points and warnings in mind:  
Infusion sets are sterile products intended for single use only. Sterility is  
guaranteed for unopened packaging up to the indicated expiration date.  
Do not use sterile products if the package is damaged.  
Properly tighten the infusion set to the adapter to avoid leakage. Turn the  
infusion set luer-lock connector until it stops. Do not turn it any further and  
do not use any auxiliary tools as this may crack the infusion set luer-lock  
connector and result in leakage.  
Your pump cannot detect infusion set leakage. You must inspect all parts  
of your infusion set at least every three hours during the day and before  
you go to sleep. If there seems to be any loss of insulin and all parts are  
correctly tightened, immediately replace the leaking component.  
Immediately check your blood glucose level, because insulin delivery has  
been interrupted. Take appropriate actions according to your doctor’s or  
healthcare team’s instructions.  
Always disconnect your infusion set before changing a cartridge or  
priming. You risk uncontrolled insulin delivery into your body. Always  
follow the instructions for the infusion set you are using.  
Have the following materials ready:  
Your ACCU-CHEK Spirit insulin pump  
A disinfectant for your skin  
A new ACCU-CHEK infusion set  
Chapter 3: Using Your Pump  
70  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Follow the steps below:  
1. Wash your hands thoroughly.  
2. Make sure your pump is in STOP mode.  
3. Remove the infusion tube set from the headset.  
4. Remove the infusion tube set from the adapter and dispose of it properly.  
Note:  
The battery key has a notch that fits  
all ACCU-CHEK infusion set luer-lock  
connectors. You may use the battery  
key to loosen the luer-lock connector  
between your infusion set and adapter  
if it cannot be loosened by hand.  
5. Carefully prepare a new ACCU-CHEK infusion set for use.  
6. Turn the ACCU-CHEK infusion set by hand clockwise into the adapter.  
7. Properly tighten the luer-lock connector by hand.  
Chapter 3: Using Your Pump  
ACCU-CHEK Spirit Insulin Pump  
71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. Prime the infusion set following all the instructions and warnings found in  
the section “Priming the Infusion Set” (page 53).  
9. Insert a new infusion set following the instructions found in the section  
“Preparing the Infusion Site and Inserting the Infusion Set” (page 56).  
10. Remove the old infusion headset and dispose of it properly.  
11. Place the pump in RUN mode. See the section “Starting Insulin Delivery”  
(page 65).  
Note:  
If you use a soft cannula-type infusion set, you must proceed with a  
bolus to fill the air space in the cannula after removing the introducer  
needle according to the instructions for the infusion set you are using.  
Failure to do so could result in missed insulin dosing. See the section  
“Programming a Standard Bolus” (page 77) for more information.  
Inspect your infusion site at least once or twice each day for irritation  
or infection. Signs of infection may include, but are not limited to, pain,  
lumps, redness, heat, or discharge.  
If you see redness or swelling, change the infusion headset and the  
infusion site immediately and contact your doctor or healthcare team.  
Change the infusion headset according to the instructions for the infusion  
set you are using and your doctor or healthcare team’s recommendations.  
Chapter 3: Using Your Pump  
72  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Changing the Cartridge and Infusion Set  
Have the following materials ready:  
Your ACCU-CHEK Spirit insulin pump  
A disinfectant for your skin  
A new ACCU-CHEK infusion set  
A filled ACCU-CHEK 3.15 ml plastic cartridge with a  
luer-lock connection  
A new adapter (must be changed at least with every 10th cartridge)  
1. Wash your hands thoroughly.  
2. Make sure your insulin pump is in STOP mode.  
3. Remove your infusion tube set from the headset.  
4. Remove the infusion tube set from the adapter and dispose of it properly.  
Note:  
The battery key has a notch that fits  
all ACCU-CHEK infusion set luer-lock  
connectors. You may use the battery  
key to loosen the luer-lock connector  
between your infusion set and adapter  
if it cannot be loosened by hand.  
Chapter 3: Using Your Pump  
ACCU-CHEK Spirit Insulin Pump  
73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Hold your ACCU-CHEK Spirit insulin pump  
with the adapter pointing downwards.  
6. Remove the adapter and cartridge from your pump by unscrewing the  
adapter from the pump casing. Pulling the adapter from the pump casing  
without completely unscrewing it may result in the plunger staying attached  
to the piston rod. If this should happen, please call Pump Support at  
1-800-688-4578.  
7. Remove the cartridge from the adapter and dispose of the cartridge properly.  
8. Hold the adapter up to the light to check for signs of wear or dirt (especially  
on the seals inside and outside of the adapter). It is recommended that you  
replace the adapter at least every 10th insulin cartridge change.  
9. If needed, clean the adapter using water and dry it. If it shows signs of  
wear or dirt, replace it immediately.  
10. Install a new cartridge and a new infusion set. See the following sections  
for instructions: “Changing the Infusion Set” (page 70), “Preparing the  
Cartridge” (page 42), “Connecting the Cartridge, Adapter, and Infusion Set”  
(page 47), “Inserting the Cartridge” (page 49), “Priming the Infusion Set”  
(page 53), “Preparing the Infusion Site and Inserting the Infusion Set”  
(page 56), and “Starting Insulin Delivery” (page 65).  
Note:  
Check the amount of insulin remaining in the cartridge at least once  
a day. Press the f key from the RUN or STOP screen to display  
the remaining cartridge content in units. Before going to sleep, make  
sure that:  
• the cartridge contains enough insulin to last through the night, and  
• that the time and date are set correctly.  
Chapter 3: Using Your Pump  
74  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4: Boluses and Temporary Basal Rates  
Understanding Boluses  
A bolus is the amount of insulin delivered (in addition to the basal rate) to cover  
the intake of food and correct high blood glucose levels. The bolus type and  
amount is determined by your doctor or healthcare team’s guidelines, your  
blood glucose level, your food intake, your health condition, and your activity  
level. Discuss the timing and the amount and type of bolus you need with your  
doctor or healthcare team. You can review the time, date, and amount of the  
last 30 boluses in the bolus history. See the section “Review the Bolus History”  
(page 107) for more information.  
Your pump can deliver three types of boluses:  
Standard Bolus  
Extended Bolus  
MultiWave Bolus  
fast delivery  
delivery over a programmed period of time  
combines fast delivery with delivery over a  
programmed period of time  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
75  
Download from Www.Somanuals.com. All Manuals Search And Download.  
• Incorrect programming of your insulin pump may cause  
inappropriate insulin delivery.  
• Your insulin pump must be programmed with your  
personal settings prior to starting insulin pump therapy.  
• Do not operate your insulin pump without knowing your  
personal settings.  
!
L
WARNING  
• If you are not sure about your personal settings or you  
are less experienced, have your doctor or healthcare  
team check your personal settings.  
• Work with your doctor or healthcare team to determine  
the timing, amount, and type of bolus you need to deliver.  
• Make sure you know your personal carbohydrate to  
insulin ratio and your correction bolus ratio.  
Chapter 4: Boluses and TBRs  
76  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Programming a Standard Bolus  
There are two ways to program a Standard Bolus:  
“Quick” Standard Bolus  
“Scroll” Standard Bolus  
Using a and s on your  
insulin pump, with bolus increments  
determined by the user  
Menu-guided using d and f  
with a and s for  
programming of the amount  
The bolus amount per delivery is limited to 25 units using U100 insulin.  
After you program a Standard Bolus there is a delay of 5 seconds before the  
bolus delivery starts. This delay lets you cancel the bolus before delivery starts  
by pressing the a or s key. Alert A8: BOLUS CANCELLED will occur. Press  
f twice to confirm and turn off the alert. See the section “Alert A8: BOLUS  
CANCELLED” (page 153) for more information.  
Programming a “Quick” Standard Bolus  
The following process involves only the a and s keys. You can start with  
either key. The recommended steps below are those described with the first  
key appearing without parenthesis, such as a. For the alternative step press  
the key in the parenthesis, such as (s).  
Note:  
Because your pump beeps and vibrates to guide you safely through the  
programming, you may program the “Quick” STANDARD Bolus using  
one finger. After the bolus is programmed, you can count the beeps and  
vibrations of your pump to verify the correct programming.  
The bolus increment for the “Quick” Standard Bolus is initially set to  
0.5 units.  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
77  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 sec.  
RUN  
screen  
bolus  
amount  
5 sec. delay  
bolus  
delivery  
if O unit  
3 sec.  
confirmation  
of amount  
5 sec. delay  
A8: bolus cancelled  
Follow the steps below:  
1. From the RUN screen press and hold s (a) until you hear a beep  
sequence and feel a vibration.  
This activates the “Quick” STANDARD  
BOLUS function.  
Hint:  
Start programming with the a key in areas of poor lighting, as this  
will also turn on the backlight.  
2. Press a (s) repeatedly until the bolus amount is reached.  
Each press of the a (s) key adds one bolus  
increment to the bolus amount. Your pump  
beeps and vibrates simultaneously once for  
each programmed bolus increment.  
Five seconds after the last press of a (s), your pump confirms the total  
bolus amount with one beep and vibration for each bolus increment programmed.  
The Standard Bolus symbol ( ) blinks for 5 seconds (the bolus delivery  
start delay).  
After this bolus delivery start delay, the pump beeps three times and then  
begins to deliver the total bolus amount programmed. The countdown of the  
remaining bolus amount appears on the display.  
Chapter 4: Boluses and TBRs  
78  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The bolus increment that can be programmed into your  
insulin pump determines the Standard Bolus amount  
using the a and s keys of your insulin pump.  
Incorrect programming of the bolus increment may  
cause inappropriate insulin delivery. See the section  
“Bolus Increment” (page 131) for more information.  
!
L
WARNING  
Cancelling a “Quick” Standard Bolus  
during programming (the bolus amount blinks):  
• Press the key s or a you used to enter the “Quick” Standard  
Bolus programming option. This will reset the bolus amount to 0.0  
units. A melody will play. If no new bolus is programmed, your pump  
returns to the RUN screen and beeps three times after five seconds.  
No bolus is delivered.  
during confirmation (when the beeps and vibrations occur), or during the  
start delay ( blinks):  
• Press s or a. A melody plays. Your pump returns to the RUN  
screen. An alert A8: BOLUS CANCELLED occurs. Press f twice to  
confirm and turn off the alert. See the section “Alert A8: BOLUS  
CANCELLED” (page 153) for more information.  
No bolus is delivered.  
during bolus delivery (countdown of bolus amount):  
• Press and hold s or a for 3 seconds until you hear a melody.  
An alert A8: BOLUS CANCELLED occurs. Press f twice to confirm  
and turn off the alert.  
Bolus delivery is interrupted.  
See the section “Alert A8: BOLUS CANCELLED” (page 153) for  
more information.  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
79  
Download from Www.Somanuals.com. All Manuals Search And Download.  
You can view the actual bolus amount delivered prior to the cancellation by  
looking at the bolus history. See the section “Review the Bolus History” (page  
107) for more information.  
Make sure that the cancellation was intended, and then program a new Standard  
Bolus, if needed.  
Programming a Menu-Guided “Scroll” Standard Bolus  
Regardless of the bolus increment setting, the bolus increments for the “Scroll”  
Standard Bolus are fixed to 0.1 unit. You can program this bolus in the STANDARD  
BOLUS screen by holding the a or s key (scrolling) until the bolus amount  
is displayed.  
Follow the steps below:  
1. Press d to move to the STANDARD BOLUS screen.  
2. Press f to select. The BOLUS AMOUNT screen appears.  
3. Press or scroll a to increase or s to decrease the bolus amount.  
4. Press f to confirm the bolus amount.  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
80  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The Standard Bolus symbol ( ) blinks for five seconds (the bolus delivery  
start delay).  
After this bolus delivery start delay, your pump beeps three times and begins to  
deliver the total bolus amount programmed. The countdown of the remaining  
bolus amount appears on the display.  
Cancelling a Menu-Guided “Scroll” Standard Bolus  
during programming (the bolus amount blinks):  
You have three options for cancelling a “Scroll” Standard Bolus during  
programming:  
• If you do not press any key for 20 seconds, your pump returns to the  
RUN screen (timeout).  
• Exit to the STANDARD BOLUS menu by pressing d+a simultaneously.  
• Set the bolus amount to 0.0 units. Press f to exit.  
No bolus is delivered.  
during the start delay ( is blinking):  
• Press s or a. A melody plays. Your pump returns to the RUN screen.  
An alert A8: BOLUS CANCELLED occurs. Press f twice to confirm  
and turn off the alert.  
No bolus is delivered.  
Please refer to section “Alert A8: BOLUS CANCELLED” (page 153) for more  
information.  
during bolus delivery (countdown of bolus amount):  
• Press and hold s or a for 3 seconds until you hear a melody.  
An alert A8: BOLUS CANCELLED occurs. Press f twice to confirm  
and turn off the alert.  
Bolus delivery is interrupted.  
Please refer to section “Alert A8: BOLUS CANCELLED” (page 153) for  
more information.  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
81  
Download from Www.Somanuals.com. All Manuals Search And Download.  
You can view the actual bolus amount delivered prior to the cancellation  
by looking at the bolus history. Please refer to the section “Review the  
Bolus History” (page 107) for more information.  
Make sure that the cancellation was intended and then program a new Standard  
Bolus, if needed.  
Chapter 4: Boluses and TBRs  
82  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Programming an Extended Bolus  
The Extended Bolus is available in the ADVANCED and CUSTOM user menus. The  
Extended Bolus function lets you program a bolus to be delivered over a period of  
time. This function is helpful during long meals, dinners, or receptions, or when  
you have meals that are digested slowly. The Extended Bolus may also be  
appropriate for people who have gastroparesis (delayed digestion).  
The duration of bolus delivery can be programmed in 15-minute intervals up to  
24 hours and begins immediately after confirmation.  
Note:  
You can add a Standard Bolus to an ongoing Extended Bolus. If you  
cancel the Standard Bolus, the Extended Bolus continues to run.  
During the delivery of an Extended Bolus, the programming of another  
Extended Bolus or a MultiWave Bolus is blocked. If you want to program  
a different Extended Bolus, put the pump in STOP mode to cancel the  
current bolus and to program a new one.  
RUN  
screen  
Extended  
bolus  
bolus  
amount  
bolus  
duration  
stop your  
ACCU-CHEK  
A8: bolus cancelled  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
83  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Follow the steps below:  
1. Press d to move to the EXTENDED BOLUS screen.  
2. Press f to select. The BOLUS AMOUNT screen appears.  
3. Press a to increase or s decrease the bolus amount.  
The duration of the last Extended Bolus  
delivered appears, or when using your pump  
for the first time, a default value appears.  
4. Press d to move to the BOLUS DURATION screen.  
Press a to increase or s decrease the  
bolus duration.  
Note:  
Press d to loop between programming the bolus amount and  
bolus duration.  
5. Press f to confirm the bolus amount and duration. You will hear a melody  
when the bolus delivery begins within the next 3 minutes.  
Chapter 4: Boluses and TBRs  
84  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
At any function screen you have three options to exit:  
If you want to deliver the bolus  
• Press f. Your pump returns to the RUN screen.  
If you want to cancel the bolus  
• Wait for your pump to return to the RUN screen (timeout), or  
• Press d+a simultaneously (exit feature) to exit the function screen.  
• Set the bolus amount to 0.0 units and press f.  
Note:  
During the entire bolus delivery, the remaining time and amount of the  
Extended Bolus and the current hourly basal rate are displayed in the  
RUN screen.  
Note:  
If a Temporary Basal Rate is active at the same time as the Extended Bolus,  
the remaining time and amount of the bolus and the amount of the  
increased or decreased hourly basal rate are displayed in the RUN screen.  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
85  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Cancelling an Extended Bolus  
during programming (the bolus amount or bolus duration blinks):  
You have three options for cancelling an Extended Bolus during  
programming:  
• If you do not press any key for 20 seconds, your pump returns to the  
RUN screen (timeout).  
• Exit to the EXTENDED BOLUS menu by pressing d+a simultaneously.  
• Set the bolus amount to 0.0 units and press f.  
No bolus is delivered.  
during bolus delivery:  
• If the Extended Bolus delivery has begun, it can be cancelled by putting  
your pump into STOP mode. This cancels the bolus delivery and an alert  
A8: BOLUS CANCELLED occurs. Press f twice to confirm and turn off  
the alert. See the section “Alert A8: BOLUS CANCELLED” (page 153) for  
more information.  
Bolus delivery is interrupted.  
The bolus amount delivered up to the cancellation can be reviewed in  
the BOLUS HISTORY screen. See the section “Review the Bolus  
History” (page 107) for more information.  
When your insulin pump is in STOP mode, it does not deliver  
any insulin. Put the pump in RUN mode to continue the  
CAUTION insulin delivery.  
!
L
Chapter 4: Boluses and TBRs  
86  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
If a Temporary Basal Rate is active at the same time, it is also cancelled  
when your pump is put into STOP mode. An alert A8: BOLUS CANCELLED  
and an alert A6: TBR CANCELLED occur.  
Press f twice to confirm and turn off the first alert. The second alert  
appears on the display. Press f twice to confirm and turn off the second  
alert. Both alerts are recorded in the alarm history.  
See the sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8:  
BOLUS CANCELLED” (page 153) for more information.  
Make sure that the cancellation was intended and then program a new  
Extended Bolus (and/or Temporary Basal Rate), if needed.  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
87  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Programming a MultiWave Bolus  
The MultiWave Bolus function is available in the ADVANCED user menu. It is  
designed to better simulate a body’s insulin delivery. It combines an immediate  
bolus delivery followed by an Extended Bolus delivery. This function can be  
helpful when you have meals that include both rapid and slowly absorbed  
carbohydrates.  
The duration of bolus delivery can be programmed in 15-minute intervals for  
up to 24 hours.  
Note:  
• The MultiWave Bolus is available in the ADVANCED user menu.  
• You can add a Standard Bolus to an ongoing MultiWave Bolus. If you  
cancel the Standard Bolus, the MultiWave Bolus remains running.  
• During the delivery of a MultiWave Bolus, the programming of a  
second MultiWave Bolus or an Extended Bolus is blocked. If you  
need a different MultiWave Bolus, put your pump in STOP mode  
to cancel the current bolus and program a new one.  
Chapter 4: Boluses and TBRs  
88  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RUN  
screen  
MultiWave  
bolus  
bolus  
amount  
bolus  
delivery  
3 sec.  
immediate  
bolus  
or  
stop your  
ACCU-CHEK  
bolus  
duration  
5 sec. delay  
A8: bolus cancelled  
Follow the steps below:  
1. Select the ADVANCED user menu or turn on the MULTIWAVE BOLUS menu  
choice in the CUSTOM user menu. Also, your pump must be in RUN mode.  
2. Press d to move to the MULTIWAVE BOLUS screen.  
3. Press f to select.  
The BOLUS AMOUNT screen appears.  
4. Press a or s to increase or decrease the total bolus amount.  
The first time you press a, the bolus duration of  
the last MultiWave Bolus appears, except when  
you are using your pump for the first time, in which  
case, a default value appears.  
As you increase or decrease the total bolus amount, the suggested immediate  
bolus amount increases or decreases proportionally.  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
89  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Press d to move to the IMMEDIATE BOLUS screen.  
Press a or s to correct and set the immediate  
bolus amount.  
6. Press d to move to the BOLUS DURATION screen.  
Press a or s to correct and set the Extended  
Bolus duration in increments of 15 minutes for up  
to 24 hours.  
Note:  
Press d to loop between programming the total bolus amount,  
immediate bolus amount, and Extended Bolus duration.  
7. Check the total and immediate bolus amounts and the extended bolus duration  
programmed on the display. The set values can be viewed in the BOLUS  
AMOUNT, IMMEDIATE BOLUS, and BOLUS DURATION screens.  
8. Press f to confirm both bolus amounts and the bolus duration.  
The MultiWave Bolus symbol ( ) blinks for 5 seconds (bolus delivery  
start delay).  
The pump beeps three times and the immediate bolus delivery begins. The  
countdown of the remaining bolus continues to appear on the display.  
Chapter 4: Boluses and TBRs  
90  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
Once the immediate bolus is delivered, the remaining time, bolus amount,  
and the current hourly basal rate are displayed on the RUN screen.  
If a Temporary Basal Rate is active at the same time, the remaining time  
and amount of the bolus and the amount of the increased or decreased  
hourly basal rate are displayed on the RUN screen.  
Cancelling a MultiWave Bolus  
during programming (a bolus amount or the bolus duration blinks):  
There are three ways to cancel a MultiWave Bolus during programming:  
• Do not press any key for 20 seconds. The pump returns to the RUN  
screen (timeout).  
• Exit to the MULTIWAVE BOLUS screen by pressing d+a simultaneously.  
• Set the total bolus amount to 0.0 units and press f.  
No bolus is delivered.  
during the start delay ( blinks):  
• Press s or a. The pump beeps and vibrates.  
The pump returns to the RUN screen.  
An alert A8: BOLUS CANCELLED occurs. Press f twice to confirm  
and turn off the alert.  
No bolus is delivered.  
See the section “Alert A8: BOLUS CANCELLED” (page 153) for more  
information  
during bolus delivery:  
You can cancel the immediate delivery by pressing and holding s or  
a for 3 seconds until you hear a melody. This cancels the whole  
bolus (immediate and Extended Bolus delivery). An alert A8: BOLUS  
CANCELLED occurs.  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
91  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
If the alert A8: BOLUS CANCELLED does not occur, the bolus delivery  
was not cancelled. Cancel the extended delivery (see below).  
• You can cancel the extended delivery by putting the pump in STOP  
mode. This cancels the Extended Bolus. An alert A8: BOLUS CANCELLED  
occurs. Press f twice to confirm and turn off the alert.  
Bolus delivery is interrupted.  
See the section “Alert A8: BOLUS CANCELLED” (page 153) for  
more information.  
You can review the actual bolus amount delivered before the cancellation  
by looking at the bolus history. See the section “Review the Bolus History”  
(page 107) for more information.  
When your insulin pump is in STOP mode, it does not deliver  
any insulin. Put your insulin pump in RUN mode to continue  
the insulin delivery.  
!
L
CAUTION  
Note:  
If a Temporary Basal Rate is active at the same time, it is also cancelled  
when your pump is put in STOP mode. An alert A8: BOLUS CANCELLED  
and an alert A6: TBR CANCELLED occur.  
Press f twice to confirm and turn off the first alert. The second alert  
appears on the display. Press f twice to confirm and turn off the second  
alert. Both alerts are recorded in the alarm history.  
See the sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8:  
BOLUS CANCELLED” (page 153) for more information.  
Make sure that the cancellation was intended and then program a new MultiWave  
Bolus (and/or Temporary Basal Rate), if needed.  
Chapter 4: Boluses and TBRs  
92  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Programming a Temporary Basal Rate (TBR)  
Your pump lets you temporarily increase or decrease your basal rate. This lets  
you match changing insulin needs from increased or decreased activity level,  
illness, or stress.  
Incorrect programming of your insulin pump may cause  
incorrect insulin delivery. Your insulin pump must be  
programmed with your personal settings prior to starting  
insulin pump therapy.  
!
L
WARNING  
Do not operate your insulin pump without knowing your  
personal settings. If you are not sure about your personal  
settings or if you are less experienced, have your doctor  
or healthcare team check your personal settings.  
Typically, a basal rate is set to 100%, but you can:  
Increase it to 200% for up to 24 hours  
Decrease it to 0% for up to 24 hours  
When you increase or decrease your basal rate, each hourly basal rate for the  
duration you program is increased or decreased. The characteristic shape of  
your basal rate profile remains and does not become a flat curve, if the change  
is larger than 0%. For 0% the basal rate profile becomes a flat curve and no  
insulin (except boluses) is delivered during the programmed duration.  
Discuss programming a Temporary Basal Rate with your doctor or healthcare team.  
Note:  
A running Temporary Basal Rate increase or decrease remains active,  
even if you changed the basal rate profile.  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
93  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RUN  
screen  
temporary  
basal rate  
precentage  
duration  
Follow the steps below:  
1. Put your pump in RUN mode.  
2. Press d to move to the TEMPORARY BASAL RATE (TBR) screen.  
3. Press f to select.  
The TBR PERCENTAGE screen appears.  
Note:  
If no Temporary Basal Rate is currently active, the percentage is set to  
100%. If a Temporary Basal Rate is active, its duration and percentage  
appear on the RUN screen.  
4. Press a or s to increase or decrease the basal rate.  
Chapter 4: Boluses and TBRs  
94  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
When you increase (or decrease) the Temporary Basal Rate the duration of  
your previous Temporary Basal Rate increase (or decrease) appears. When  
you use your pump for the first time, a default value appears.  
5. Press d to move to the TBR DURATION screen.  
Press a or s to correct and set the duration.  
Note:  
Press d to loop between programming the basal rate percentage  
and duration.  
6. Press f to save and exit.  
The new Temporary Basal Rate is immediately activated.  
Note:  
At any function screen you have three options to exit:  
if you want to save the changes  
• Press f. Your pump returns to the RUN screen.  
if you want to undo the changes  
• Wait for your pump to return to the RUN screen (timeout), or  
• Press d+a simultaneously (exit feature) to exit the function screen.  
• Set volume to 100% and press f.  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
95  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
During the Temporary Basal Rate delivery, the increase or decrease in  
percentage, the increased or decreased hourly basal rate (hourly basal rate  
plus TBR percentage), and the remaining time are displayed in the RUN  
screen. An arrow pointing up symbolizes a Temporary Basal Rate increase;  
an arrow pointing down symbolizes a Temporary Basal Rate decrease.  
Note:  
If an Extended Bolus or a MultiWave Bolus is active at the same time, the  
remaining time and amount of the bolus and the amount of the increased  
or decreased hourly basal rate are displayed in the RUN screen.  
TBR OVER Alert  
At the end of a Temporary Basal Rate, an alert A7: TBR OVER occurs. Press f  
twice to confirm and turn off the alert. See the section “Alert A7: TBR OVER”  
(page 153) for more information.  
Chapter 4: Boluses and TBRs  
96  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Cancelling a Temporary Basal Rate  
during programming:  
There are three ways to cancel a Temporary Basal Rate during programming:  
• Do not press any key for 20 seconds. The pump returns to the  
RUN screen (timeout).  
• Exit to the TEMPORARY BASAL RATE (TBR) screen by pressing d+a  
simultaneously.  
• Set the TBR percentage to 100%. Press f to save and exit.  
No Temporary Basal Rate is delivered.  
during delivery:  
There are two ways to cancel a Temporary Basal Rate during delivery:  
• Set the TBR percentage to 100%:  
1. Press d to move to the TEMPORARY BASAL RATE (TBR) screen.  
2. Press f to select.  
The TBR PERCENTAGE screen appears.  
3. Press a or s to return the basal rate to 100%.  
Press f to save and exit.  
• Or, put your pump in STOP mode.  
The Temporary Basal Rate delivery is interrupted.  
An alert A6: TBR CANCELLED occurs. Press f twice to confirm and  
turn off the alert.  
Chapter 4: Boluses and TBRs  
ACCU-CHEK Spirit Insulin Pump  
97  
Download from Www.Somanuals.com. All Manuals Search And Download.  
When your insulin pump is in STOP mode, it does not deliver  
any insulin. Put your insulin pump in RUN mode to continue  
the insulin delivery.  
!
L
CAUTION  
Note:  
If an Extended Bolus or a MultiWave Bolus is active at the same time,  
it is also cancelled when your pump is put into STOP mode. An alert  
A8: BOLUS CANCELLED and an alert A6: TBR CANCELLED occur.  
Press f twice to confirm and turn off the first alert. The second alert  
appears on the display. Press f twice to confirm and turn off the  
second alert. Both alerts are recorded in the alarm history.  
See the sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8:  
BOLUS CANCELLED” (page 153) for more information.  
Make sure that the cancellation was intended and then program a new  
Temporary Basal Rate (and/or an Extended or MultiWave Bolus), if needed.  
Chapter 4: Boluses and TBRs  
98  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5: User Menus  
You will need to access various screens so you can view stored information,  
program the pump, perform a function, or change a setting. Your pump offers  
a choice of three different user menus:  
• The STANDARD User Menu  
This menu offers all the functions you need for successful insulin pump  
therapy and lets you change between the user menus. If you are new  
to pump therapy, start with this menu. Later, as your comfort with  
pump therapy grows, you may choose the features offered in the  
ADVANCED user menu.  
• The ADVANCED User Menu  
This menu offers the full range of the ACCU-CHEK Spirit insulin pump  
functions. It includes all functions from the STANDARD user menu, plus  
a wide range of functions for the more experienced user.  
• The CUSTOM User Menu  
Because the ACCU-CHEK Spirit insulin pump has many features, you  
may want to choose the screens that you see when you scroll through  
the menu choices. This menu lets you display or hide other screens.  
Note:  
At the end of this user guide you will find diagrams for the STANDARD  
and ADVANCED user menus.  
For your safety and convenience during programming, your insulin pump  
returns itself to the RUN or STOP screen if no key is pressed within 20  
seconds. For your safety, the changes you made before your insulin  
pump “times out” will not be saved.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Selecting a User Menu  
RUN  
screen  
select the  
user menu  
user  
menu  
select basal  
rate profile  
Note:  
When you change the user menu, your current basal rate profile will not  
appear if it is not activated. Make sure that your desired basal rate  
profile number(s) is/are activated. This can be done through the previous  
user menu or the ADVANCED user menu.  
Follow the steps below:  
1. Press d to move to the SELECT USER MENU screen.  
Press f to select.  
Chapter 5: User Menus  
100  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. The current user menu is selected on the SELECT USER MENU screen.  
Press a or s to select a different user menu.  
3. Press f to confirm.  
The selected basal rate profile and its daily insulin  
total appear,  
or  
if the previous basal rate profile is not available  
from the selected user menu, dashes appear instead.  
4. Press a or s to select a basal rate profile, if needed.  
Check the (new) basal rate profile and its daily insulin total.  
5. Press f to save and exit. The new user menu and basal rate profile are  
active immediately.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
101  
Download from Www.Somanuals.com. All Manuals Search And Download.  
A running increased or decreased Temporary Basal  
Rate will remain active, even if you changed the basal  
rate profile. The temporary basal rate is based on the  
new basal rate.  
!
L
CAUTION  
Note:  
At any function screen you have three options to exit:  
If you want to confirm and save the changes press f until your  
pump returns to the RUN screen,  
If you want to undo the changes  
press d+a simultaneously to exit the function screen, or  
Wait for your pump to return to the RUN screen (timeout).  
Chapter 5: User Menus  
102  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
STANDARD User Menu  
The following is a list of menu choices for the STANDARD user menu as they  
are addressed in the manual. To view the order in which they appear on the  
pump, refer to the menu diagrams at the back of the manual. Some of the  
functions have already been described earlier in the manual. The remaining  
functions are described following this list.  
SET TIME AND DATESee the section “Setting Time and Date” (page 32)  
PROGRAM BASAL RATE PROFILE - See the section “Programming your  
Basal Rate Profile” (page 37)  
CHANGE THE CARTRIDGESee the section “Inserting the Cartridge” (page 49)  
PRIME THE INFUSION SETSee the section “Priming the Infusion Set”  
(page 53)  
START YOUR ACCU-CHEKSee the section “Starting Insulin Delivery”  
(page 65)  
STOP YOUR ACCU-CHEKSee the section “Stopping Insulin Delivery”  
(page 66)  
STANDARD BOLUSSee the section “Programming a Standard Bolus”  
(page 77)  
EXTENDED BOLUSSee the section “Programming an Extended Bolus”  
(page 83)  
TEMPORARY BASAL RATE (TBR)See the section “Programming a  
Temporary Basal Rate (TBR)” (page 93)  
COMMUNICATIONSee the section “Data Transfer” (page 104)  
INFORMATIONSee the section “Reviewing Pump Data” (page 107)  
SETUP MENU STANDARDSee the section “SETUP MENU Standard”  
(page 113)  
SELECT USER MENUSee the section “Selecting a User Menu” (page 100)  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
103  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Data Transfer  
The pump memory stores up to 4,500 pump events (about 90 days of usage).  
Pump events include alerts and errors, programming operations, and insulin  
delivery records. Some or all of the pump memory may be transferable to  
compatible software.  
The built-in infrared interface inside the bottom cover of your insulin pump  
allows for wireless data transfer.  
Chapter 5: User Menus  
104  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Setup for Data Transfer  
STOP  
screen  
communi-  
cation  
start data  
transfer  
stop data transfer  
Follow the steps below:  
1. Put your pump in STOP mode.  
2. Press d to move to the COMMUNICATION screen.  
3. Press f. The DATA TRANSFER screen appears and a melody signals that  
your pump is ready to deliver data.  
When the data transfer is finished, press f to return to the STOP screen.  
Set up your pump with a cartridge, adapter, and a new infusion set and put  
it in the RUN mode, if necessary.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
105  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
Your pump returns to STOP mode 15 minutes after the last data transfer  
when the DATA TRANSFER screen is selected.  
You may press d+a simultaneously to exit while data is being  
transferred. The data transfer is interrupted and an error E12: DATA  
INTERRUPTED occurs. See the section, “Error E12: DATA INTERRUPTED”  
(page 163) for more information.  
Chapter 5: User Menus  
106  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Reviewing Pump Data  
The pump memory lets you review specific data such as:  
• Bolus history  
last 30 boluses  
• Alarm history  
last 30 alerts and errors  
• History of daily insulin totals  
• Temporary Basal Rate history  
• Maintenance (pump timer)  
last 30 daily totals of insulin delivered  
last 30 increases and decreases  
remaining time in days until the  
pump timer expires  
Review the Bolus History  
The bolus history lets you review up to the last 30 boluses delivered beginning  
with the most recent entry and working backward in time.  
Each BOLUS HISTORY screen displays the:  
• bolus type ( Standard Bolus, Extended Bolus or MultiWave Bolus),  
• bolus duration ( only displayed if an Extended or a MultiWave Bolus),  
• time (  
)
• date ( ) and  
• entry number (01 is the most recent) of total entries (example 01/30) of  
a delivered bolus.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
107  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RUN  
screen  
information  
menu  
bolus  
history  
alarm  
his
daily  
tot
TBR  
his
1. Press d to move to the INFORMATION screen.  
2. Press f to select.  
The BOLUS HISTORY screen appears. The most  
recent bolus information is displayed.  
3. Press a or s to review the entries.  
Press f to exit.  
Chapter 5: User Menus  
108  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Review the Alarm History  
The alarm history lets you review the last 30 alerts or errors, beginning with  
the most recent entry and working backward in time.  
Each alarm history screen displays the:  
• number and type of the alert or error (example A6),  
• alert or error (example TBR CANCELLED),  
• time ( ),  
• date ( ) and  
• entry number (01 is the most recent) of total entries (example: 01/30)  
of an alert or error.  
1. Press d to move to the INFORMATION screen.  
Press f to select.  
2. Press d to move to the ALARM HISTORY screen.  
3. Press a or s to review the entries.  
Press f to exit.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
109  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Review the History of Daily Insulin Totals  
This history lets you review the last 30 daily totals of insulin delivered (from  
midnight to midnight; basal rate plus boluses), starting with the most recent  
entry and working backward in time.  
Each daily insulin total screen displays the following history:  
• daily total of insulin delivered ( ),  
• its date ( ), and  
• entry number (01 is the most recent) of total entries (example 01/30)  
of a daily insulin total.  
Follow the steps below:  
1. Press d to move to the INFORMATION screen.  
Press f to select.  
2. Press d to move to the DAILY TOTALS screen.  
3. Press a or s to review the entries.  
Press f to exit.  
Chapter 5: User Menus  
110  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Review the Temporary Basal Rate History  
The Temporary Basal Rate (TBR) history lets you review the last 30 TBR  
increases or decreases starting with the most recent entry and working  
backward in time.  
Each TBR history screen displays the:  
• TBR increase ( ) or decrease ( ) in percentage,  
• TBR duration ( ),  
• time (  
)
• date ( ) and  
• entry number (01 is the most recent) of total entries (example 01/30)  
of a delivered Temporary Basal Rate.  
Follow the steps below:  
1. Press d to move to the INFORMATION screen.  
Press f to select.  
2. Press d to move to the TBR HISTORY screen.  
3. Press or scroll a or s to review the entries.  
Press f to exit.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
111  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Review the Time Remaining  
Your pump is designed to be a highly reliable system that will give you worry-free  
performance for a long time. For best performance, a pump timer limits the  
operating time of the pump. This timer counts the number of days of operation  
remaining for the ACCU-CHEK Spirit insulin pump.  
Before the pump timer expires, an alert will remind you of the upcoming end of  
operating time so that you can arrange to get a new one. The pump timer symbol  
(
) appears in the RUN screen or STOP screen as a reminder. When the timer  
expires, an error occurs and your pump goes into STOP mode. It can no longer  
be put into RUN mode. See the sections “Alert A5: PUMP TIMER” (page 151)  
and “Error E5: END OF OPERATION” (page 158) for more information.  
Review the number of days left until the pump timer expires in the INFORMATION  
menu.  
Follow the steps below:  
1. Press d to move to the INFORMATION screen.  
Press f to select.  
2. Press d to move to the PUMP TIMER screen.  
The days until the pump timer expires is shown.  
3. Press f to exit.  
Chapter 5: User Menus  
112  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SETUP MENU Standard  
The SETUP MENU Standard should not be confused with the STANDARD user  
menu. The SETUP MENU Standard is a menu choice that appears in STANDARD  
and ADVANCED user menus. The SETUP MENU Standard allows you to “set up”  
certain features and settings. The following features and settings can be  
changed using the SETUP MENU Standard menu choice.  
Turning KeyLock On or Off  
The KeyLock function lets you lock the four keys of your pump and serves as an  
extra safety measure against activating functions by accident (for example, while  
sleeping, during contact sports). To use the KeyLock function, it must be turned on.  
See the section “Keys and Key Combinations” (page 11) for more information.  
RUN / STOP  
screen  
setup  
standard  
KeyLock  
beep  
vo
alarm  
sig
automatic  
off  
battery  
typ
orientation  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
113  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Follow the steps below:  
1. Press d to move to the SETUP MENU STANDARD screen.  
Press f to select.  
2. The current KeyLock status (ON or OFF) appears.  
3. Press a or s to turn the KeyLock ON or OFF.  
Press f to save and exit.  
If no key is pressed for 10 seconds after your pump has returned to the RUN  
screen or STOP screen and if the KeyLock is turned on, all four keys are locked.  
Chapter 5: User Menus  
114  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Unlocking the KeyLock  
3 sec.  
RUN screen  
locked  
10 sec.  
RUN screen  
unlocked  
Follow the steps below:  
1. From the RUN or the STOP screen, press d+s simultaneously (you hear  
three beeps in RUN mode one beep in STOP mode) and hold for 3 seconds  
until you again hear three beeps (in RUN mode) one beep (in STOP mode) to  
confirm that the KeyLock is unlocked.  
2. If no key is pressed for 10 seconds after your pump has returned to the RUN  
or the STOP screen, all four keys are automatically locked.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
115  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Adjusting the Beep Volume  
Your pump beeps when a key is pressed or when an alert or error occurs.  
Follow the steps below:  
1. Press d to move to the SETUP MENU STANDARD screen.  
Press f to select.  
2. Press d to move to the BEEP VOLUME screen.  
The active beep volume is displayed.  
3. Press a or s to adjust the beep volume.  
There are five volume levels:  
no beeps (beeps are turned off)  
very low  
low  
normal  
maximum  
4. Press f to save and exit.  
Note:  
The volume of the STOP-Warning is independent of the programmed  
beep volume. It always occurs at the maximum level. When the beeps  
are turned off and an alert or error occurs, the beeps will become active  
again after a short period of time to make sure that you are aware of  
the alarm or error. These beeps will steadily increase in volume until  
the maximum level is reached if the alert or error is not cancelled.  
Chapter 5: User Menus  
116  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Setting Up Alarm Signals  
The ACCU-CHEK Spirit insulin pump lets you choose how you want your pump  
to alert you when an alarm occurs.  
You have three options:  
beep only  
vibrate only  
beep and vibrate  
Follow the steps below:  
1. Press d to move to the SETUP MENU STANDARD screen.  
Press f to select.  
2. Press d to move to the ALARM SIGNALS screen.  
The active alarm signals are displayed.  
3. Press a or s to set the alarm signals.  
4. Press f to save and exit.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
117  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Automatic Off  
Discuss using the automatic off function with your doctor  
or healthcare team.  
!
L
CAUTION  
Automatic off is a safety feature that stops insulin delivery and triggers an  
error E3: AUTOMATIC OFF if no keys are pressed within a programmed time  
period in RUN mode.  
The automatic off can either be set to OFF or programmed up to 24 hours in  
1-hour intervals in the SETUP MENU STANDARD screen.  
Follow the steps below:  
1. Press d to move to the SETUP MENU STANDARD screen.  
Press f to select.  
2. Press d to move to the AUTOMATIC OFF screen.  
The current status is displayed at the bottom  
left of the screen.  
3. Press or scroll a to increase or s to decrease the duration for the  
automatic off in 1-hour intervals.  
4. If you would like to set the automatic off function to OFF, press or scroll s  
until 0 hours and OFF appears on the display.  
5. Press f to save and exit.  
Chapter 5: User Menus  
118  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Battery Type  
If you do not set the correct battery type on your insulin  
pump, the alert A2: BATTERY LOW may not occur in time  
to provide you sufficient warning to replace your battery.  
!
L
WARNING  
Your pump lets you choose between using 1.5 volt AA alkaline batteries or  
rechargeable AA NiMH batteries. If you change from one battery type to the  
other, you must also change the battery type setting.  
See the section “Accessories and DisposablesBattery” (page 21) for more  
information on recommended batteries.  
Follow the steps below:  
1. Press d to move to the SETUP MENU STANDARD screen.  
Press f to select.  
2. Press d to move to the BATTERY TYPE screen.  
The current battery type is displayed.  
3. Press a or s to select the battery type ALKALINE or NiMH  
(rechargeable) battery.  
Press f to save and exit.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
119  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Display Orientation  
Your pump lets you flip the display orientation by 180° so that you can view it  
more conveniently depending on how you are wearing the pump.  
Follow the steps below:  
1. Press d to move to the SETUP MENU STANDARD screen.  
Press f to select.  
2. Press d to move to the ORIENTATION screen.  
3. Press a or s to select the display orientation.  
4. Press f to save and exit.  
Chapter 5: User Menus  
120  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
If you flip the display orientation of your insulin pump  
by 180°, the a and s keys will also reverse their  
functions. In comparison to the display orientation, the  
upper key will be the a and the lower key will be s.  
This change applies for all a and s functions including  
turning on the backlight. The d and f will not change  
in their function, but will remain the same, regardless of  
your screen orientation.  
!
L
CAUTION  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
121  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ADVANCED User Menu  
The ADVANCED user menu offers the full range of the ACCU-CHEK Spirit  
insulin pump functions. The ADVANCED user menu includes all functions  
from the STANDARD user menu, plus a wide range of functions for the more  
experienced user.  
The following is a list of additional menu choices that appear in the ADVANCED  
user menu. Some of the functions have already been discussed earlier in the  
manual. The remaining functions are described below.  
EXTENDED BOLUS–See the section “Programming an Extended Bolus”  
(page 83)  
MULTIWAVE BOLUSSee the section “Programming a MultiWave Bolus”  
(page 88)  
CHANGE BASAL RATE PROFILESee the section “Selecting a Basal  
Rate Profile” (page 125)  
PROGRAM BASAL RATE PROFILESee the section “Programming  
Additional Basal Sate Profiles” (page 124)  
ALARM CLOCKSee the section “Alarm Clock” (page 127)  
SETUP MENU ADVANCEDSee the section “SETUP MENU Advanced”  
(page 129)  
Chapter 5: User Menus  
122  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basal Rate Profiles  
Your pump can store up to five different basal rate profiles to match your  
changing insulin needs (for example, Monday-Friday versus Exercise Day  
versus Sleeping Late Day).  
Discuss additional basal rate profiles with your doctor or healthcare team.  
Consult with your doctor or healthcare team before changing basal rate profiles,  
as this may change how you use your pump.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
123  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Programming Additional Basal Rate Profiles  
Incorrect programming of your insulin pump may cause  
!
incorrect insulin delivery. Your insulin pump must be  
programmed with your personal settings prior to starting  
insulin pump therapy. Do not operate your insulin pump  
without knowing your personal settings. If you are not  
sure about your personal settings or you are less  
experienced, have your doctor or healthcare team  
check your personal settings.  
L
WARNING  
Follow the steps below:  
1. Press d to move to the PROGRAM BASAL RATE PROFILE screen.  
Note:  
Only basal rate profiles that are accessible in your current user menu  
are available.  
2. The hourly basal rates for each additional basal rate profile are set in the  
same way as for basal rate profile 1. See the section “Programming your  
Basal Rate Profile” (page 37) for more information.  
Chapter 5: User Menus  
124  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Selecting a Basal Rate Profile  
RUN  
screen  
change the  
BR profile  
BR profile  
no. …  
Follow the steps below:  
1. Press d to move to the CHANGE BASAL RATE PROFILE screen.  
2. Press f to select.  
The current basal rate profile is displayed.  
Note:  
Only basal rate profiles that are accessible in your current user menu  
are available.  
3. Press a or s to select a basal rate profile. The selected basal rate profile  
and its daily insulin total appear.  
4. Press f to select. The new basal rate profile is active immediately.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
125  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note:  
When you change your user menu your current basal rate profile may  
not appear.  
If the basal rate profile number(s) you want does not appear,  
• Select the ADVANCED user menu, or  
• Program the basal rate profile to an available basal rate profile number  
A running Temporary Basal Rate increase or decrease stays active,  
even if you change the basal rate profile on your pump.  
Chapter 5: User Menus  
126  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alarm Clock  
You can set the alarm clock for a single alarm or for multiple alarms. Multiple  
alarms repeat at the set time every day. Alarms can be set to remind you of  
blood glucose testing or other important events.  
Set Single or Multiple Alarms  
1. Press d to move to the ALARM CLOCK screen.  
2. Press f to select. The current alarm clock status (OFF, ONCE, or EVERY  
DAY) and alarm time appear.  
3. Press a or s to change the alarm clock to OFF, ONCE ( ), or EVERY  
DAY ( ).  
4. Press d to move to the SET ALARM HOUR screen.  
Press a or s to set the hour.  
5. Press d to move to the SET ALARM MINUTE screen.  
Press a or s to set the minute.  
6. Press f to save and exit.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
127  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Turning Off the Alarm Clock  
When the alarm clock goes off, an alert A4: ALARM CLOCK occurs. Press f  
twice to confirm and turn off the alert.  
See the section “Alert A4: ALARM CLOCK” (page 151) for more information.  
Chapter 5: User Menus  
128  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SETUP MENU Advanced  
The SETUP MENU Advanced should not be confused with the ADVANCED user  
menu. The SETUP MENU Advanced is a menu choice that appears in the  
ADVANCED user menu. The SETUP MENU Advanced allows you to “set up” certain  
features and settings. The following features and settings can be changed using  
the SETUP MENU Advanced menu choice.  
Time Format  
You can set the time in either of the following formats:  
American:  
12-hour clock (AM/PM), example 1:39 PM  
European/Military: 24-hour clock (00:00-23:59), example 13:39  
Follow the steps below:  
1. Press d to move to the SETUP MENU ADVANCED screen.  
Press f to select.  
2. The TIME FORMAT screen appears.  
Press a or s to select a time format.  
3. Press f to save and exit.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
129  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Date Format  
You can set the date in the following formats:  
American:  
European:  
mm/dd/yyyy, example 04/26/06  
dd.mm.yyyy, example 26.04.06  
Follow the steps below:  
1. Press d to move to the SETUP MENU ADVANCED screen.  
Press f to select.  
2. Press d to move to the DATE FORMAT screen.  
Press a or s to select a date format.  
3. Press f to save and exit.  
Chapter 5: User Menus  
130  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Bolus increment  
The bolus increment that can be programmed into your  
!
insulin pump determines the “Quick” Standard Bolus  
amount using the a and s keys of your insulin  
pump. Incorrect programming of the bolus increment  
may cause inappropriate insulin delivery.  
L
WARNING  
To address patients who may need large or small amounts of insulin, the bolus  
increment for the ACCU-CHEK Spirit insulin pump is adjustable. The “Quick”  
Standard Bolus is initially set to 0.5 units per key press but can be changed  
in the SETUP MENU ADVANCED to 0.1, 0.2, 0.5, 1.0, or 2.0 units. Units for the  
scroll “Standard” bolus, which may be adjusted by simply holding down the  
a or s keys, will move at a constant 0.1 units. See the section  
“Programming a Standard Bolus” (page 77) for more information.  
Follow the steps below:  
1. Press d to move to the SETUP MENU ADVANCED screen.  
Press f to select.  
2. Press d to move to the BOLUS INCREMENT screen.  
Press a or s to select a bolus increment.  
3. Press f to save and exit.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
131  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Prime Quantity  
The prime quantity needed to fill an infusion set depends on the length of the  
infusion set tubing being used. The shorter your infusion set tubing, the less  
insulin will be required to prime the infusion set. The default setting for the  
prime quantity is 25 units of U100 insulin.  
Note:  
The amount of insulin used for the priming is not added to the history of  
daily insulin totals. You can press any of your pump’s keys to stop priming.  
Follow the steps below:  
1. Press d to move to the SETUP MENU ADVANCED screen.  
Press f to select.  
2. Press d to move to the PRIME QUANTITY screen.  
Press a or s to select a prime quantity  
between 0.0 and 30.0 units.  
3. Press f to save and exit.  
Chapter 5: User Menus  
132  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Lock Basal Rate Profiles  
Your pump lets you change your hourly basal rates as needed. You may lock  
your basal rate profiles to provide extra protection against accidentally changing  
a basal rate during normal operation. When this function is activated, your hourly  
basal rates cannot be changed.  
Follow the steps below:  
1. Press d to move to the SETUP MENU ADVANCED screen.  
Press f to select.  
2. Press d to move to the BASAL RATE LOCK screen.  
Press a or s to turn the basal rate lock ON  
or OFF.  
3. Press f to save and exit.  
Note:  
If the basal rate lock is set to ON (  
), the programming of the basal  
rate profiles 1, 2, 3, 4, and 5 is blocked. To check if the basal rate lock  
is ON, press d to move to the PROGRAM BASAL RATE screen and  
press f to select. The  
indicates the lock of the basal rate programming. (The  
symbol in the BASAL RATE TOTAL screen  
symbol appears  
in the BASAL RATE TOTAL screen only.) Press d to view the hourly  
basal rates. The a and s functions are blocked.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
133  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Language  
Your pump is programmed in several languages. These languages include  
Italian, English, German, French, Spanish, Portuguese, Czech, Dutch, Swedish,  
Norwegian, Polish, and Greek.  
To choose a language, do the following:  
1. Press d to move to SETUP MENU ADVANCED screen.  
Press f to select.  
2. Press d to move to the LANGUAGE screen.  
Press a or s to select a language.  
3. Press f to save and exit.  
Chapter 5: User Menus  
134  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Display Contrast  
Your pump lets you adjust the display contrast.  
Follow the steps below:  
1. Press d to move to the SETUP MENU ADVANCED screen.  
Press f to select.  
2. Press d to move to the DISPLAY CONTRAST screen.  
Press a or s to select a contrast level.  
3. Press f to save and exit.  
Chapter 5: User Menus  
ACCU-CHEK Spirit Insulin Pump  
135  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CUSTOM User Menu  
The CUSTOM user menu contains the 11 most commonly used functions according  
to a customer survey.  
The following is a list of the functions included in the CUSTOM user menu. All  
of these functions have been addressed in detail in STANDARD and ADVANCED  
user menu sections.  
– STANDARD BOLUS  
– EXTENDED BOLUS  
– TEMPORARY BASAL RATE (TBR)  
– INFORMATION  
– CHANGE BASAL RATE PROFILE  
– PROGRAM BASAL RATE PROFILE 1  
– PROGRAM BASAL RATE PROFILE 2  
– ALARM CLOCK  
– SET DATE AND TIME  
– SETUP MENU STANDARD  
– SELECT USER MENU  
Chapter 5: User Menus  
136  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6: Care and Maintenance  
Care of Your ACCU-CHEK Spirit Insulin Pump  
Do not perform any servicing or repair on your insulin  
!
pump. For servicing or repairs contact Pump Support at  
1-800-688-4578.  
L
CAUTION  
Disetronic Medical Systems, as the manufacturer, guarantees the functions of  
your pump according to its specifications only if all servicing of your pump is  
done and/or authorized by Disetronic Medical Systems.  
System Inspection Checklist  
Your pump must be well maintained to guarantee accurate insulin delivery.  
Check the pump display at least every three hours during the day and before  
you go to sleep, and especially if for any reason you might not be able to  
hear the beeps or to feel the vibrations. Use only sterile products and  
accessories designed to be used with your pump. Replace and discard these items  
according to your doctor’s or healthcare team’s recommendations and the instruc-  
tions that come with these items.  
Chapter 6: Care and Maintenance  
ACCU-CHEK Spirit Insulin Pump  
137  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Check your pump daily. Make sure that:  
• The pump casing, the display, and the cartridge are not chipped or  
cracked and that the display doesn’t show incomplete or abnormal  
letters or symbols.  
• You visually inspect the cartridge. Make sure that the actual amount of  
insulin in the cartridge equals the displayed cartridge amount on the  
QUICK INFO screen.  
You inspect all parts of your pump and infusion set at least every  
three hours during the day and before you go to sleep. If you detect  
any loss of insulin, immediately replace the leaking component.  
• The battery cover is correctly tightened and positioned even with the  
pump casing.  
• The adapter is inserted properly.  
• Your infusion set is primed, free of air bubbles, and tight in the adapter.  
• Your infusion set is inserted according to the instructions for its use.  
• Your infusion site is secure, comfortable, and free of irritation or infection.  
• The cartridge is free of air bubbles.  
• Your pump is in RUN mode.  
• The basal rates are set correctly according to your doctor’s or healthcare  
team’s recommendations.  
• The time and date are set correctly.  
• The Temporary Basal Rate changes are set according to your doctor’s  
or healthcare team’s recommendations.  
• The beeps and/or vibrations are set.  
• You have your personal emergency kit.  
Chapter 6: Care and Maintenance  
138  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
What to Do When You Drop Your Pump  
If the pump is dropped it could be harmed or the seal that protects it against water  
could be damaged. Be careful not to drop your pump. Use the proper ACCU-CHEK  
Spirit insulin pump carrying system to help avoid dropping your pump. See  
Disetronic Medical Systems sterile products and accessories catalogs and  
brochures for more information. Or, contact Pump Support at 1-800-688-4578.  
If your pump is dropped:  
• Check that all connections in your infusion set are still tight, and if  
they are not, reconnect or tighten them  
• Check your pump, its sterile products, and accessories for chips and  
cracks, and  
• Change the cartridge if it is damaged.  
Check your insulin pump and its sterile products and  
accessories at least every three hours. Immediately inspect  
your insulin pump if it is dropped. Do not use your insulin  
pump if chips or cracks are visible. Chips and cracks may  
allow water, dust, insulin, or other foreign substances to  
enter your insulin pump and lead to malfunction.  
!
L
CAUTION  
If you have questions or are not sure whether your pump has been damaged,  
contact Pump Support at 1-800-688-4578.  
Chapter 6: Care and Maintenance  
ACCU-CHEK Spirit Insulin Pump  
139  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Storing Your ACCU-CHEK Spirit Insulin Pump  
When your insulin pump will not be in use for a long period of  
time, it should be properly stored, to prevent a malfunction.  
!
L
CAUTION  
To store your pump, place in STOP mode, then:  
• Remove the battery to preserve the battery life  
• Remove the cartridge  
• Insert the battery cover and the adapter  
• Store your pump in its shipping case  
Storage Conditions  
Temperature  
Air humidity  
Barometric pressure  
41°F to 113°F (+5°C to +45°C)  
5 to 85% relative air humidity  
700 to 1060 mbar  
Disposal  
Your ACCU-CHEK Spirit Insulin Pump  
If you need to dispose of your insulin pump, return it to Disetronic Medical  
Systems for professional disposal.  
Battery  
Properly dispose of dead batteries.  
Cartridge, Infusion Set, Adapter, and Accessories  
Dispose of these items in your normal garbage. Remember to dispose of the  
introducer needle safely (needle container). To prevent injury to others, replace  
the needle protective cap or insert the needle of the infusion set into a disposable  
object, or put it in a hard shell container.  
Chapter 6: Care and Maintenance  
140  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance  
Cleaning Your ACCU-CHEK Spirit Insulin Pump  
The best time to clean your pump is when you change the cartridge. Use a  
soft, dry cloth for cleaning. If the pump compartments are very dirty, contact  
Pump Support at 1-800-688-4578 for more instructions.  
Always remove the cartridge and adapter and place your  
insulin pump in STOP mode while cleaning. Avoid pressing  
!
L
the keys of your insulin pump during cleaning, as this may  
accidentally change your settings. Do not use alcohol,  
solvents, strong detergents, bleaching agents, scouring  
pads, or sharp instruments for cleaning as they may  
damage the pump.  
CAUTION  
Battery Information  
It is important to change your battery regularly and to have spare batteries ready.  
The life of any battery is affected by how you use your insulin pump, your personal  
settings, delivery rates, temperature, and other factors. With typical pump use  
[50 U/day using U100 insulin; operating temperature 72°F 6°F (22°C 3°C)],  
pump battery life is about four weeks using alkaline batteries and one week  
using rechargeable batteries. See the section “Accessories and Disposables–  
Battery” (page 21) for more information on recommended batteries. See the  
section “Inserting (or Changing) the Battery” (page 25) for instructions on how to  
change your battery.  
Chapter 6: Care and Maintenance  
ACCU-CHEK Spirit Insulin Pump  
141  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Repair  
Do not perform any servicing or repair on your insulin pump.  
Do not use any lubricants on the pump mechanism. If you  
have questions, contact Pump Support at 1-800-688-4578.  
!
L
CAUTION  
If you have problems with your pump, contact Pump Support at 1-800-688-4578.  
Pump Support can help you and answer many questions over the telephone.  
Pump Support may direct you to return your pump to Disetronic Medical Systems  
for repair. For example, you might need to return your pump if an alert or error  
could not be cleared by following the procedures described in “Alerts and Errors.”  
(page 147). If Pump Support directs you to return your pump, they will give you  
a Return Material Authorization (RMA) number that you must include with the  
pump when you send it.  
If you must ship your pump, carefully pack the pump and the used cartridge,  
battery, battery cover, adapter, and infusion set to avoid damage to them during  
transportation. Place your pump back in the shipping case and put the case in a  
package that will not rip or tear. It is best to send your pump using a shipping  
method and carrier that tracks the shipment.  
When you return your pump, include the following:  
• a description of the reason for returning your pump  
• your name and address  
• your daytime phone  
• the serial number of your pump, and  
• any RMA (Return Material Authorization) number received from  
Pump Support.  
By returning your pump to Disetronic Medical Systems, you allow Disetronic  
Medical Systems to undertake any testing on your infusion pump that is needed for  
a proper inspection, and understand that your pump may not be returned to you.  
Chapter 6: Care and Maintenance  
142  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Troubleshooting  
This section describes problems that may occur while using your ACCU-CHEK  
Spirit insulin pump. See “Alerts and Errors” (page 147) for more information.  
Note:  
The solutions in this guide are only suggestions. Do not use these  
procedures without approval of your doctor or healthcare team. Always  
follow your doctor’s or healthcare team’s recommendations, and  
contact them if you have any concerns about your insulin pump therapy.  
You need to know that the interruption of insulin delivery (due to a leak,  
occlusion, or loss of insulin potency, for example) or the malfunction of  
your insulin pump can result in a rapid rise of your blood glucose level.  
Although your pump has an internal security check system, it cannot  
alert you if your infusion set is leaking or if the insulin you are using  
has lost its potency.  
Chapter 6: Care and Maintenance  
ACCU-CHEK Spirit Insulin Pump  
143  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Therapy-Related Issues  
Many factors, such as alcohol consumption, medications other than insulin,  
ineffective or expired insulin, decreased activity, illness, and stress can affect  
blood glucose levels. Contact your doctor or healthcare team for advice on how  
to manage these or any other therapy related issues.  
Troubleshooting Table  
Problem  
Recommended Actions  
Individual characters, numbers, or Contact Pump Support at  
symbols on the display are shown 1-800-688-4578.  
incompletely or not at all.  
Your insulin pump does not beep Check that the beep volume and alarm  
or vibrate.  
signals are turned on. Check and  
change the battery. If your insulin pump  
still does not beep or vibrate, contact  
Pump Support at 1-800-688-4578.  
Self-test incomplete (after an error Remove the battery, wait a few  
E7: ELECTRONIC ERROR).  
seconds, and insert it again. If the  
self-test remains incomplete, contact  
Pump Support at 1-800-688-4578.  
Incorrect insulin delivery due to  
incorrect time setting.  
The correct hourly basal rate cannot be  
delivered if the time is not set correctly.  
Check and reset the time if it is wrong.  
Do the same check when you change  
the battery.  
Incorrect insulin delivery due to  
Always review changes made in basal  
incorrect basal rate programming. rate programming. Confirm that you are  
using your correct basal rate profile.  
Chapter 6: Care and Maintenance  
144  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Problem  
Recommended Actions  
Air bubbles in the cartridge  
and/or infusion set  
Check blood glucose level. In the  
cartridge: Disconnect the infusion tube  
set from the headset and re-prime your  
infusion tube set. If the air bubbles  
remain, replace the cartridge. Make sure  
to fill the cartridge with insulin at room  
temperature. In the infusion set:  
Disconnect and prime your infusion set.  
Reconnect and make sure all connections  
are tight and secure.  
Blood in the infusion set tubing Change your infusion set and infusion site.  
Empty cartridge  
Replace the cartridge. Cartridges are for  
single use only.  
Disconnected or dislodged  
infusion set  
Make sure all connections are tight. Inspect  
connections for leaking insulin. Check your  
blood glucose level. Change your infusion  
set and infusion site.  
Leak in system  
Check for leaking insulin at all connections  
and on the skin. Change your infusion set,  
infusion site, cartridge, and adapter. Check  
your blood glucose level.  
Chapter 6: Care and Maintenance  
ACCU-CHEK Spirit Insulin Pump  
145  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Problem  
Recommended Actions  
Occlusion in system  
Disconnect and re-prime your infusion  
set. Check your blood glucose. If the  
occlusion remains, prime the pump  
without the cartridge inserted. See the  
section “Error E4: OCCLUSION” (page 156).  
Failure to complete prime  
Poor absorption of insulin  
Re-prime the infusion set. Always prime  
until a bubble-free insulin flow emerges  
from the tip of the needle. Make sure the  
tubing is free of air.  
Choose infusion sites that do not have  
scar tissue, bruises, or tissue build-up.  
Do not use the same infusion set or  
infusion site longer than your doctor or  
healthcare team recommends.  
Infusion site becomes sore,  
red, or swollen.  
Change your infusion headset and infusion  
site immediately. Use proper insertion  
technique. Check blood glucose level.  
Follow the planned infusion site rotation  
pattern and advice given by your doctor or  
healthcare team.  
Chapter 6: Care and Maintenance  
146  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7: Alerts and Errors  
A safety system monitors and controls the function of your pump. This safety  
system performs more than 9 million safety checks every day. If it detects a  
change from the pump’s normal state, it will trigger an alert (warning instruction)  
or an error (error message). Beeps, vibrations, and messages on the pump  
display indicate the cause of an alert or error. Alerts and errors are the only  
way you will be quickly notified of any changes in your pump.  
Check the pump display at least every three hours during  
the day, and before you go to sleep, and especially if you  
might be not be able to hear the beeps or feel the vibrations.  
!
L
CAUTION  
Your pump gives alerts and errors using beeps and vibrations. You can turn off  
either the beeps or the vibrations, but for your safety you cannot turn off both  
at the same time for the initial alarm signal.  
If you do not confirm an alert or error within 60 seconds, it will signal again.  
When the beep is turned off or programmed at lower than maximum volume  
level, the beep volume will rise to the loudest volume level if not confirmed.  
The beep and vibration will continue at this level until confirmed.  
Note:  
The volume of the STOP-Warning is separate from the programmed beep  
volume and cannot be changed. It always occurs at the maximum level.  
KeyLock is unlocked by the system each time an alert or error occurs.  
Chapter 7: Alerts and Errors  
ACCU-CHEK Spirit Insulin Pump  
147  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Confirming an Alert or Error  
When you confirm A1: CARTRIDGE LOW  
, E1: CARTRIDGE EMPTY,  
A2: BATTERY LOW , E2: BATTERY DEPLETED, A5: PUMP TIMER  
or E5:  
END OF OPERATION, the symbol for that alert or error stays in the RUN screen  
or STOP screen as a reminder. For other alerts or errors, alarm information is  
cleared from the display but stays in the alarm history and the event memory.  
Follow the steps below to confirm an alert or error:  
1. Press f to turn off the beeps and vibrations.  
The alert or error code stays on the display.  
2. As soon as you understand the reason for the alert, error code, or message,  
confirm by pressing f again.  
3. Take the appropriate action, if any.  
4. If more than one alert and/or error occurs simultaneously, press f twice  
for each alert or error to confirm and turn each off.  
5. After an alert or error, make sure that your pump is returned to RUN mode.  
Chapter 7: Alerts and Errors  
148  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List of Alerts  
After an alert occurs, your insulin pump may be in STOP  
!
mode and the insulin delivery may be interrupted. In order  
to maintain insulin delivery, you must act immediately  
according to the instructions given for each error code  
or put your insulin pump in RUN mode to continue the  
insulin delivery.  
L
WARNING  
Alert A1: CARTRIDGE LOW  
The contents of the insulin cartridge have dropped to 20 units of insulin. Be  
prepared to replace the cartridge as soon as possible. The cartridge low symbol  
(
) appears in the RUN or STOP screen as a reminder until you change  
the cartridge.  
1. Press f twice to turn off and to confirm the alert.  
Change the cartridge before it is empty.  
2. If necessary, put your pump in RUN mode.  
See the section “Changing the Cartridge and Infusion Set” (page 73) for more  
information.  
Chapter 7: Alerts and Errors  
ACCU-CHEK Spirit Insulin Pump  
149  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alert A2: BATTERY LOW  
The battery power is low. Be prepared for a battery change as quickly as possible.  
The battery low symbol (  
) appears in the RUN or STOP screen as a reminder  
until you change the battery.  
1. Press f twice to turn off and to confirm the alert.  
Replace the battery as soon as possible (within a  
few hours). Failure to change the battery within a  
few hours of receiving the alert could result in  
additional error messages.  
2. If necessary, put your pump in RUN mode.  
See the section “Accessories and DisposablesBattery” (page 21) for more  
information on recommended batteries. See the section “Inserting (or Changing)  
the Battery” (page 25) for instructions on how to change your battery.  
Alert A3: REVIEW TIME AND DATE  
The pump was without power for more than one hour. The ACCU-CHEK Spirit  
insulin pump settings and history data are saved in memory regardless of  
power loss. However, it may be necessary to enter the time and date again.  
1. Press f twice to turn off and to confirm the alert.  
Review the time and date and correct them,  
if they are wrong.  
2. If necessary, put your pump in RUN mode.  
See the section “Setting Time and Date” (page 32) for more information.  
Chapter 7: Alerts and Errors  
150  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alert A4: ALARM CLOCK  
An alarm clock reminder was programmed into the pump.  
1. Press f twice to turn off and to confirm the alert.  
2. If necessary, put your pump in RUN mode.  
Note:  
If you have set the alarm occurrence to EVERY DAY, the same alert occurs  
every 24 hours. Turn the alarm clock function OFF if you do not want  
the alarm to go off again at that same time every day.  
See the section “Alarm Clock” (page 127) for more information.  
Alert A5: PUMP TIMER  
Your pump’s operating time will expire soon. It is recommended that you replace  
your pump after a period of four years of running time. After the alert A5: PUMP  
TIMER occurs, the ACCU-CHEK Spirit insulin pump will display the symbol  
remind you to check the timer.  
to  
1. Press f twice to turn off and to confirm the alert.  
2. Put your pump in RUN mode.  
3. Go to the INFORMATION screen on your pump.  
Chapter 7: Alerts and Errors  
ACCU-CHEK Spirit Insulin Pump  
151  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Move to the PUMP TIMER screen.  
The days until the pump timer expires are shown.  
When the PUMP TIMER reaches zero, the pump will change to STOP mode and  
no longer operate. An error, E5: END OF OPERATION, will appear on the display.  
Contact Pump Support at 1-800-688-4578 to discuss your options for further  
use of your ACCU-CHEK Spirit insulin pump.  
Note:  
The alert A5: PUMP TIMER ALERT and the error E5: END OF OPERATION  
are based on running time (the total time the pump is in RUN mode)  
and not calendar years. However, the warranty is based on the date the  
pump was purchased.  
See the section “Review the Time Remaining” (page 112) for more information.  
Alert A6: TBR CANCELLED (Temporary Basal Rate Cancelled)  
A temporary increase or decrease of the basal rate has been cancelled.  
1. Press f twice to turn off and to confirm the alert.  
2. If necessary, put your pump in RUN mode.  
3. Make sure that the cancellation was intended and then program a new  
Temporary Basal Rate if needed.  
Chapter 7: Alerts and Errors  
152  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alert A7: TBR OVER (Temporary Basal Rate Over)  
A Temporary Basal Rate has ended. The original basal rate (100%) will resume  
automatically once the alert has been confirmed.  
1. Press f twice to turn off and confirm the alert.  
2. Decide whether a further Temporary Basal Rate change is needed. If necessary,  
program a new Temporary Basal Rate.  
Alert A8: BOLUS CANCELLED  
A bolus was cancelled during the start delay or after delivery began.  
1. Press f twice to turn off and to confirm the alert.  
2. If necessary, put your pump in RUN mode.  
3. Make sure that the cancellation was intended and then program a new  
bolus if needed.  
Note:  
The actual bolus amount that was delivered before the cancellation can be  
reviewed in the bolus history. See the section “Review the Bolus  
History” (page 107) for more information.  
Chapter 7: Alerts and Errors  
ACCU-CHEK Spirit Insulin Pump  
153  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List of Errors  
After the occurrence of an error, your insulin pump will be in  
!
STOP mode and the insulin delivery is interrupted. In order  
to maintain insulin delivery, you must act immediately  
according to the instructions given for each error code.  
L
WARNING  
Error E1: CARTRIDGE EMPTY  
The insulin cartridge is empty. Change the cartridge immediately.  
1. Press f twice to turn off and to confirm the error.  
2. Disconnect or remove the infusion set from your infusion site.  
3. Change the cartridge and prime the infusion set.  
4. Reconnect the infusion set at your infusion site.  
5. Put your pump in RUN mode.  
See the section “Changing the Cartridge and Infusion Set” (page 73) for more  
information.  
Chapter 7: Alerts and Errors  
154  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Error E2: BATTERY DEPLETED  
The battery is out of power. Replace it immediately.  
After inserting a new battery, always check that the time  
!
and date of your insulin pump are set correctly. Incorrect  
programming of the time and date may cause incorrect  
insulin delivery. If you, your doctor, or healthcare team  
store and analyze your therapy data electronically and the  
time and date of the devices used are not set identically,  
the gathered data might not be meaningful.  
L
WARNING  
1. Press f twice to turn off and to confirm the error.  
2. Replace the battery.  
3. Put your pump in RUN mode.  
Note:  
Your pump stores the settings you saved before the battery power ran  
out. If your pump is without power for more than one hour, check the  
time and date.  
See the section “Accessories and DisposablesBattery” (page 21) for more  
information on recommended batteries. See the section “Inserting (or changing)  
the Battery” (page 25) for instructions on how to change your battery.  
Chapter 7: Alerts and Errors  
ACCU-CHEK Spirit Insulin Pump  
155  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Error E3: AUTOMATIC OFF  
Automatic off is a safety feature that stops insulin delivery and triggers an  
error E3: AUTOMATIC OFF if no keys are pressed within a programmed time  
period in RUN mode.  
1. Press f twice to turn off and to confirm the error.  
2. Put your pump in RUN mode.  
Error E4: OCCLUSION  
If an error E4: OCCLUSION occurs, immediately check your  
blood glucose level, and check again in about 2 hours,  
because insulin delivery has been interrupted. If your blood  
glucose level is high, take appropriate actions according to  
your doctor’s or healthcare team’s instructions.  
!
L
WARNING  
Insulin (a maximum of 3.5 units of U100 insulin) was not delivered. This may  
be due to:  
• a blocked infusion set,  
• a reused cartridge, or  
• a dirty or damaged piston rod.  
Note:  
The display of this error depends on several factors including your current  
hourly basal rate and boluses. For example, a blocked infusion set with  
a basal rate of 2.0 U/h (U100 insulin and no additional boluses) would  
trigger an error E4: OCCLUSION in approximately three hours or less,  
depending on the type and severity of the blockage.  
Chapter 7: Alerts and Errors  
156  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
If the occlusion (blockage) was inside your infusion set, follow the  
steps below.  
1. Press f twice to turn off and to confirm the error.  
2. Disconnect or remove the infusion set from your infusion site.  
3. Change the infusion set and prime it.  
4. Reconnect the infusion set (tubing) at your infusion site.  
5. Put your pump in RUN mode.  
In rare cases, the cartridge itself can be the cause of the occlusion. If an  
error E4: OCCLUSION occurs again, you must change the cartridge. Follow  
the steps below:  
1. Press f twice to turn off and to confirm the error.  
2. Disconnect or remove the infusion set from your infusion site.  
3. Change the cartridge and prime the infusion set.  
4. Reconnect the infusion set (tubing) at your infusion site.  
5. Put your pump in RUN mode.  
Chapter 7: Alerts and Errors  
ACCU-CHEK Spirit Insulin Pump  
157  
Download from Www.Somanuals.com. All Manuals Search And Download.  
If the error E4: OCCLUSION occurs again, prime your pump without a  
cartridge. Follow the steps below:  
1. Press f twice to turn off and to confirm the error.  
2. Disconnect or remove the infusion set from your infusion site.  
3. Remove the cartridge, adapter and infusion set from your pump.  
4. Carry out the PRIME THE INFUSION SET function without the cartridge,  
adapter, and infusion set inserted.  
5. Insert a new cartridge with connected adapter and infusion set.  
6. Prime the infusion set.  
7. Reconnect the infusion set (tubing) at your infusion site.  
8. Put your pump in RUN mode.  
If the error E4: OCCLUSION continues to occur, contact your doctor or  
healthcare team for an alternate therapy plan. Contact Pump Support at  
1-800-688-4578 for more assistance.  
Error E5: END OF OPERATION  
Your pump timer has reached zero. The pump will change to STOP mode and will  
no longer operate. An error E5: END OF OPERATION will appear on the display. It  
is recommended that you replace your pump after four years. Before this error  
occurs, an alert A5: PUMP TIMER will remind you to replace your pump soon.  
1. Press f twice to turn off and to confirm the error.  
Chapter 7: Alerts and Errors  
158  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Contact Pump Support at 1-800-688-4578 to discuss your options for further  
use of the ACCU-CHEK Spirit insulin pump.  
Note:  
The alert A5: PUMP TIMER ALERT and the error E5: END OF OPERATION  
are based on running time (the total time the pump is in RUN mode)  
and not calendar years. However, the warranty is based on the date the  
pump was purchased.  
Error E6: MECHANICAL ERROR  
A mechanical error was discovered during a routine automatic system check  
and insulin delivery has stopped.  
1. Press f twice to turn off and to confirm the error.  
2. Remove or disconnect the infusion set from your infusion site.  
3. Remove the cartridge, adapter, infusion set (tubing), and battery from  
your pump. Insert a new battery.  
4. Rewind the piston rod.  
5. Install a new cartridge and infusion set.  
6. Prime the infusion set.  
7. Reconnect the infusion set (tubing) at your infusion site.  
8. Put your pump in RUN mode.  
Chapter 7: Alerts and Errors  
ACCU-CHEK Spirit Insulin Pump  
159  
Download from Www.Somanuals.com. All Manuals Search And Download.  
If the error E6: MECHANICAL ERROR occurs again in spite of these actions,  
contact your doctor or healthcare team for an alternate therapy plan. Contact  
Pump Support at 1-800-688-4578 for more assistance.  
Error E7: ELECTRONIC ERROR  
An electronic error was discovered during a routine automatic system check  
and the insulin delivery has stopped. You cannot turn off and confirm an error  
E7: ELECTRONIC ERROR by pressing f. You must remove the battery instead.  
1. Remove or disconnect the infusion set from your infusion site.  
2. Insert a new battery into the pump.  
3. Prime the infusion set.  
4. Reconnect the infusion set (tubing) at your infusion site.  
5. Put your pump in RUN mode.  
If an error E7: ELECTRONIC ERROR occurs during programming (i.e., basal rate,  
time and date) check your settings.  
If the error E7: ELECTRONIC ERROR occurs again, contact Pump Support at  
1-800-688-4578 for assistance. Contact your doctor or healthcare team if  
you need an alternate therapy plan.  
Chapter 7: Alerts and Errors  
160  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Error E8: POWER INTERRUPT  
A power interrupt error occurs when you insert a new battery without first placing  
the pump in STOP mode. See the section “Inserting (or Changing) the Battery”  
(page 25) for instructions on how to change your battery. The error can also occur  
if the pump was dropped. If the pump was dropped, see the section “What to Do  
When You Drop Your Pump” (page 139) for more information.  
1. Press f twice to turn off and to confirm the error.  
2. Your pump goes in STOP mode.  
3. Check the time and date and correct them if they are wrong. See the section  
“Setting Time and Date” (page 32) for more information.  
4. If a bolus and/or a Temporary Basal Rate was interrupted by the error,  
review the bolus and/or the Temporary Basal Rate history for the delivered  
amount and duration. See the section “Reviewing Pump Data” (page 107)  
for instructions.  
5. Put your pump in RUN mode.  
6. Program a new bolus and/or Temporary Basal Rate if you need them. See  
“Boluses and Temporary Basal Rates” (page 75).  
If an error E8: POWER INTERRUPT occurs during programming, check your settings.  
Chapter 7: Alerts and Errors  
ACCU-CHEK Spirit Insulin Pump  
161  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Error E10: CARTRIDGE ERROR  
The CHANGE THE CARTRIDGE function was not correctly carried out.  
1. Press f twice to turn off and to confirm the error.  
2. Remove or disconnect the infusion set from your infusion site.  
3. Start the CHANGE THE CARTRIDGE menu.  
4. Prime the infusion set.  
5. Reconnect the infusion set (tubing) at your infusion site.  
6. Put your pump into RUN mode.  
7. If an error E10: CARTRIDGE ERROR occurs during the winding of the piston  
rod, insert a new battery into the pump and repeat steps 1-6.  
Error E11: SET NOT PRIMED  
Never prime an infusion set that is connected to your body.  
You risk uncontrolled insulin delivery into your body. With  
detachable infusion sets, make sure you have disconnected  
the tubing from your site prior to priming. Always follow  
the instructions provided with your infusion set.  
!
L
WARNING  
The cartridge was changed but the infusion set was not primed. The error will  
display when you try to put your pump in RUN mode.  
Chapter 7: Alerts and Errors  
162  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. Press f twice to turn off and to confirm the error.  
2. Remove or disconnect the infusion set from your infusion site.  
3. Prime the infusion set.  
4. Reconnect the infusion set (tubing) at your infusion site.  
5. Put your pump in RUN mode.  
See the sections “Changing the Cartridge and Infusion Set” (page 73) and  
“Priming the Infusion Set” (page 53) for more information.  
Error E12: DATA INTERRUPTED  
If data transfer between compatible software and your  
insulin pump is disrupted in any way, the configuration  
may be incomplete and an error E12: DATA INTERRUPTED  
occurs. The data transfer must be completed successfully  
before you can put your insulin pump into RUN mode.  
!
L
CAUTION  
Data transfer between your pump and a PC was interrupted before it was  
completed.  
1. Press f twice to turn off and to confirm the error.  
2. Restart the data transfer.  
Before you can put your pump in RUN mode, the data transfer must be successful.  
Chapter 7: Alerts and Errors  
ACCU-CHEK Spirit Insulin Pump  
163  
Download from Www.Somanuals.com. All Manuals Search And Download.  
See the section “Data Transfer” (page 104) for more information.  
Error E13: LANGUAGE ERROR  
The language setting needs to be checked.  
1. Press f twice to turn off and to confirm the error.  
2. Follow the steps in the section “SETUP MENU AdvancedLanguage”  
(page 134) to change the language.  
3. Put your pump in RUN mode.  
Chapter 7: Alerts and Errors  
164  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendixes  
Appendix A: Technical Data  
General Technical Data  
Air humidity  
• During operation: 20 to 90%  
relative humidity  
• Stored in its shipping case:  
5 to 85% relative humidity  
Barometric pressure  
During operation or stored in its  
shipping case: 700 to 1,060  
mbar [do not exceed 10,000 feet  
(3,000 meters) above sea level].  
Basal rate  
Battery life  
Minimum 0.1 U/h; maximum 25 U/h  
There are 24 hourly basal rates  
adjustable in 0.1 unit increments.  
If used in a typical usage pattern  
[50 U/day using U100 insulin;  
operating temp. of 72°F 6°F  
(22°C 3°C)] the battery life is  
approx. four weeks for alkaline  
and one week for rechargeable  
batteries.  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
165  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Bolus  
The max. bolus amount per delivery  
is 25 units when using U100 insulin.  
The bolus amount for the “Quick”  
Standard Bolus is adjustable in  
increments of 0.1, 0.2, 0.5, 1.0,  
and 2.0 units. For the “Scroll”  
Standard Bolus, the Extended  
Bolus, and the MultiWave Bolus,  
the amount is adjustable in fixed  
increments of 0.1 units. The  
duration of the Extended Bolus and  
the MultiWave Bolus is adjustable in  
intervals of 15 minutes (15 minutes  
up to 24 hours).  
Cartridge  
ACCU-CHEK 3.15 ml plastic  
cartridge with luer-lock connection.  
Data storage time  
The time and date is safely stored  
in the memory for about 1 hour  
after the battery has been removed.  
Your insulin pump’s settings (the  
hourly basal rates, remaining  
cartridge content, bolus increments,  
and active user menu) and the event  
memory (bolus history, history of  
daily insulin totals, Temporary  
Basal Rate history, alarm history)  
are saved, regardless of battery  
voltage and time the insulin pump  
has been without a battery.  
Data transfer  
Delivery  
Infrared interface  
1/20th of the current hourly basal  
rate in 3-minute intervals.  
Dimensions (maximum)  
without adapter  
Approx. 3.2 x 2.2 x 0.8 inches  
(81 x 55 x 20 mm)  
Appendixes  
166  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Flow (delivery rate)  
0.2 U/sec during infusion set  
priming and bolus when using  
U100 insulin.  
Infusion sets  
IPX 8  
ACCU-CHEK infusion sets with  
luer-lock connector.  
Protected against the effects of  
temporary immersion in water  
under standardized conditions  
[up to 60 minutes and 8 feet  
(2.5 meters)].  
Maximum time for an error  
E4: OCCLUSION  
Plastic cartridges*  
• at med. basal rate 1.0 U/h using  
U100 insulin  
5h  
• at min. basal rate 0.1 U/h using  
U100 insulin  
U100 insulin  
50h  
3.5 U  
Maximum volume* before error  
Maximum pressure  
400kPa (4.0 bar)  
U100 insulin  
Maximum quantity delivered at  
a single fault condition  
1.0 U  
Power supply  
One 1.5 volt AA Alkaline battery  
or one rechargeable Nickel Metal  
Hydride (NiMH) AA battery. Alkaline  
batteries should have a minimum  
capacity of 2500 mAh and NiMH  
AA rechargeable batteries a  
minimum capacity of 1500 mAh.  
Use only a battery charger  
officially recommended by the  
battery manufacturer.  
Pump casing  
Shock, scratch, pharmaceutical  
resistant plastic. All edges are  
rounded.  
*Determined using measurement methods according to IEC 60601-2-24-1998.  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
167  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Safety system  
Alert system, beeps, information  
on the display, vibrations, dual  
microprocessor. The insulin pump is  
controlled by two microprocessors.  
The safety concept is based on one  
processor (supervisor processor)  
supervising the other (main  
processor). Whenever a defect or  
fault occurs in the main processor,  
as identified by the supervisor  
processor, the motor is immediately  
switched off and an error E7:  
ELECTRONIC ERROR occurs.  
Conversely, the main processor  
can also determine at any time  
whether the supervisor is working  
correctly. The brushless motor also  
constitutes an important safety  
component. It provides the best  
possible reliability and accuracy in  
insulin delivery.  
Temperature ranges  
Temporary Basal Rate  
• During operation: 41°F to 104°F  
(5°C to 40° C)  
• Stored in its shipping case:  
41°F to 113°F (5°C to 45° C)  
Adjustable in 10% increments,  
0-90% for decreases, 110-200%  
for increases. The duration is  
adjustable in 15-minute intervals,  
up to a maximum of 24 hours. The  
last programmed duration is given  
by default for the next temporary  
basal rate change programmed.  
Weight  
• Empty: approx. 2.8 oz. (80 g)  
• Pump w/battery, full plastic  
cartridge, and infusion set:  
approx. 4 oz. (110 g)  
Appendixes  
168  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Technical Standards on Electromagnetic Emissions  
The standard concerning electromagnetic compatibility of medical equipment  
(IEC 60601-1-2) requires the concrete specification of the corresponding levels  
which refer to specified electromagnetic interferences. The following is  
Guidance and manufacturer’s declaration on electromagnetic emissions.  
Emissions Test Compliance Electromagnetic Environment – Guidance  
RF emissions  
CISPR11  
Group 1  
The ACCU-CHEK Spirit insulin pump uses RF  
energy only for its internal function.  
Therefore, its RF emissions are very low  
and are not likely to cause any interference  
in nearby electronic equipment.  
RF emissions  
CISPR11  
Class B  
The ACCU-CHEK Spirit insulin pump is  
suitable for use in establishments, including  
domestic establishments and those directly  
connected to the public low-voltage power  
supply network that supplies buildings used  
for domestic purposes.  
Harmonic  
emissions  
IEC 61000-3-2  
Not applicable  
Not applicable  
Voltage  
fluctuations/  
flicker emissions  
IEC 61000-3-3  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
169  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note 1  
Note 2  
At 80 and 800 MHz, the higher frequency range applies.  
These guidelines may not apply in all situations. Electromagnetic  
propagation is affected by absorption and reflection from structures,  
objects and people.  
a
Field strengths from fixed transmitters, such as base stations for radio  
(cellular/cordless) telephones and land mobile radios, amateur radio,  
AM and FM radio broadcast and TV broadcast, cannot be predicted  
theoretically with accuracy. To assess the electromagnetic environment  
due to fixed RF transmitters, an electromagnetic site survey should be  
considered. If the measured field strength in the location in which the  
ACCU-CHEK Spirit insulin pump is used exceeds the applicable RF  
compliance level above, the ACCU-CHEK Spirit insulin pump should be  
observed to verify normal operation. If abnormal performance is  
observed, additional measures may be necessary, such as reorienting  
or relocating the ACCU-CHEK Spirit insulin pump.  
b
Over the frequency range 150 kHz to 80 MHz, field strengths  
should be less than 10 V/m.  
V1 in V:  
Rated maximum 150 kHz to 80 MHz  
output power out of ISM (Industrial,  
of transmitter W Scientific, Medical Band)  
Separation distance according to frequency of transmitter m  
10.00  
E1 in V/m:  
80 MHz to  
800 MHz  
10.00  
80 MHz to  
2.5 GHz  
0.01  
0.1  
1
10  
100  
0.04  
0.11  
0.35  
1.11  
3.50  
0.04  
0.11  
0.35  
1.11  
3.50  
0.07  
0.22  
0.70  
2.21  
7.00  
Appendixes  
170  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Technical Standards on Electromagnetic Immunity  
Guidance and manufacturer’s declaration – electromagnetic immunity  
The ACCU-CHEK Spirit insulin pump is intended for use in the electromagnetic  
environment specified below. The customer or the user of the ACCU-CHEK  
Spirit insulin pump should assure that it is used in such an environment.  
IEC 60601  
Immunity Test Test Level  
Compliance  
Level  
Electromagnetic  
Environment – Guidance  
Electrostatic  
discharge (ESD)  
IEC 61000-4-2  
6 kV contact 8 kV contact  
8 kV air 15 kV air  
Avoid any contact with  
synthetic materials. A  
(IEC 60601-2-24) relative humidity of 10% will  
allow ESD to be more likely.  
Power frequency 3 A/m 400 A/m 400 A/m  
(50/60Hz)  
magnetic field  
Power frequency magnetic  
fields should be at levels  
characteristic of a typical  
IEC 61000-4-8  
(IEC 60601-2-24) location in a typical  
commercial or hospital  
environment.  
Conducted RF 3 Vrms  
IEC 61000-4-8 150 kHz to  
80 MHz  
Not applicable  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
171  
Download from Www.Somanuals.com. All Manuals Search And Download.  
IEC 60601 Compliance  
Immunity Test Test Level Level  
Radiated RF 3 V/m 10 V/m  
Electromagnetic  
Environment – Guidance  
Portable and mobile RF  
communications equipment  
should be used no closer to  
any part of the ACCU-CHEK  
Spirit insulin pump, includ-  
IEC 61000-4-3 80 MHz to 80 MHz  
2.5 GHz 2.5 GHz  
(IEC 60601-2-24) ing cables, than the recom-  
mended separation distance  
calculated from the equation  
applicable to the frequency  
of the transmitter.  
Recommended separation  
distance:  
Ҁ
.
d = 0.4 ǰP 80 MHz  
to 800 MHz  
Ҁ
.
d = 0.7 ǰP 800 MHz  
to 2.5 GHz  
where P is the maximum  
output power rating of the  
transmitter in watts (W)  
according to the transmitter  
manufacturer and d is the  
recommended separation  
distance in meters (m).  
Field strengths from fixed RF  
transmitters, as determined  
by an electromagnetic site  
survey (a), should be less  
than the compliance level in  
each frequency range (b).  
Interface may occur in the  
vicinity of equipment  
marked with the following  
symbol:  
Appendixes  
172  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Accuracy of Flow Rate for Plastic Cartridges  
Trumpet curve plotted from data after the end of the stabilization period  
The trumpet curve shows the accuracy of the delivery rate in relation to the  
observation period.  
The measurements were made at a medium basal rate of 1.0 U/h according to  
IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room  
temperature.  
The average deviation of the delivered amount (overall mean percentage flow  
error) for U100 insulin is 5%.  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
173  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Start-up graph over the stabilization period  
The start-up graph shows changes in the flow rate over the stabilization time.  
The measurements were made at a medium basal rate of 1.0 U/h according to  
IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room  
temperature.  
Influence of height on delivery accuracy  
The average deviation of the delivered amount (overall mean percentage flow  
error) for U100 insulin is 10% when your ACCU-CHEK Spirit insulin pump is  
located 3.3 feet (1 m) above/below the level of infusion site.  
Appendixes  
174  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Trumpet curve plotted from data after the end of the stabilization period  
The trumpet curve shows the accuracy of the delivery rate in relation to the  
observation period.  
The measurements were made at a medium basal rate of 0.1 U/h according to  
IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room  
temperature.  
The average deviation of the delivered amount (overall mean percentage flow  
error) for U100 insulin is 20%.  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
175  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Start-up graph over the stabilization period  
The start-up graph shows changes in the flow rate over the stabilization time.  
The measurements were made at a medium basal rate of 0.1 U/h according to  
IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room  
temperature.  
Bolus  
For U100 insulin the delivery accuracy of a maximum bolus is 5% and the  
delivery accuracy of a minimum bolus is 30%.  
The measurements were made according to IEC 60601-2-24:1998 with a  
Disetronic Classic infusion set PC 16/110 at room temperature.  
Delivery Mode  
Delivery Rate/Volume  
Accuracy  
Range  
Basal  
Bolus  
0.1 U/h  
0.1 U  
20%  
30%  
0.08 - 0.12 U/h  
0.07 - 0.13 U  
Appendixes  
176  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix B: Configuration Parameters  
Your pump is built by Disetronic Medical Systems with a standard configuration  
(factory settings). A list containing the important factory settings is added to the  
original insulin pump package when your pump is first shipped. You may need  
to adapt the factory settings to meet your individual needs. Set the configuration  
parameters directly on your insulin pump. Always consult your doctor or  
healthcare team before changing parameters.  
The list on the next two pages gives you the full range for the adjustable  
configuration parameters on your pump. Also included are examples of typical  
range and parameter settings.  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
177  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Typical Range and  
Parameter Settings  
Maximum Parameter  
Range  
Basal Rate  
– hourly basal rate +  
max. TBR  
U100: 62.5 IU/h Insulin Type Max. Amount  
U100 0 – 62.50 IU/h  
(hourly basal rate  
combined with temporary  
basal rate, max. increase)  
– hourly basal rates  
increment  
0.1 U  
– hourly basal rates  
U100: 10.0 IU/h Insulin Type Max. Amount  
U100 0 – 25.0 IU/h  
– temporary basal rate  
– TBR increase or decrease  
in increments of  
– TBR duration  
– TBR duration increments  
100%  
0 – 200%  
10%  
max. 24 h  
15 min  
10%  
15 min – 24 h  
15 min, 30 min, 1 h  
– basal rate profiles available  
– basal rate lock  
5
off  
1 – 5  
on or off  
Bolus  
– bolus amount per  
bolus delivery  
U100: 25.0 IU/h Insulin Type Max. Amount  
U100 0 – 25.0 U  
– bolus amount increments  
for “Quick” and “Scroll”  
Standard Bolus  
0.5U  
0.1, 0.2, 0.5, 1.0 or 2.0 U  
– bolus duration (Extended  
bolus, MultiWave bolus)  
– bolus duration increments  
max. 24 h  
15 min  
15 min – 24 h  
15 min, 30 min, 1h  
Appendixes  
178  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Typical Range and  
Parameter Settings  
Maximum Parameter  
Range  
Prime volume  
U100: 25.0 U  
Insulin Type  
U100  
Max. Amount  
0 – 25.0 IU/h  
Display orientation  
Display contrast  
standard  
medium  
standard or inverted  
13 steps  
User menus  
STANDARD  
3 (STANDARD, ADVANCED  
or CUSTOM)  
Volume (for beeps)  
maximum  
off, 1 – 4 steps  
Alarm signals turned on beeps and vibrations  
beeps, vibrations or both  
Battery type  
AA battery  
AA Alkaline or  
AA NiMH rechargeable battery  
KeyLock  
off  
off  
turned on or off  
Automatic off  
turned on or off  
0.1 – 24h  
Alarm clock  
off  
turned on or off  
when on: once or every  
day’s occurrence  
Time format  
Date format  
American (12 h am/pm) or  
European (24 h)  
American (mm/dd/yy) or  
European (dd.mm.yy)  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
179  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix C: Abbreviations  
approx.  
BR  
Approximately  
Basal rate  
BRP  
DDM  
h
Basal rate profile  
Diabetes Data Management  
Hour(s)  
IEC  
International Electrotechnical Commission  
incl.  
IR  
Inclusive  
Infrared  
I.U., IU  
International Units in context with biological effectiveness of  
a certain insulin amount  
kPa  
LCD  
LED  
Kilo pascals  
Liquid Crystal Display  
Light Emitting Diode for infrared communication with, for  
example, a personal computer (PC)  
NiMH  
rel.  
Nickel Metal Hydride (used with rechargeable batteries)  
Relative  
sec.  
Second(s)  
TBR  
Temporary Basal Rate  
U
see I.U.  
U/h  
Amount of International Units of insulin delivered per hour  
Each milliliter of liquid contains 100 International Units of insulin  
American date format: month/day/year  
European date format: day.month.year  
U100  
mm/dd/yy  
dd.mm.yy  
Appendixes  
180  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix D: Beeps and Melodies  
You can turn off either the pump beeps or the vibrations, but not both  
simultaneously. Please refer to sections “Adjust the Beep Volume”(page 116),  
and “Setting Up Alarm Signals” (page 117) for more information.  
Each press of  
Each press of  
Each press of  
confirms each programmed bolus  
increment when programming a “Quick” Standard Bolus  
Each press of confirms the STOP-Warning  
Temporary Basal Rate is active  
Unlock the KeyLock in RUN mode, enter the RUN or INFO  
screen, start a Temporary Basal Rate  
Lock the KeyLock in RUN mode  
Unlock the KeyLock in STOP mode, enter STOP screen  
Lock the KeyLock in STOP mode  
Exit a screen  
Cancel a running bolus or stop priming  
Start-up procedure successful  
Copy an hourly basal rate  
Alert and error  
Maximum amount reached  
Minimum amount reached  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
181  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix E: Vibrations  
Your pump signals alerts and errors using beeps and vibrations. You can turn off  
either the beeps or the vibrations, but you cannot turn off both simultaneously.  
When you program a “Quick” Standard Bolus, your ACCU-CHEK Spirit insulin pump  
confirms all programming by vibrating. You cannot turn off these vibrations.  
Your ACCU-CHEK Spirit insulin pump vibrates when  
l (Short vibration)  
Your ACCU-CHEK Spirit insulin pump performs a  
start-up procedure  
(Medium vibration)  
Your ACCU-CHEK Spirit insulin pump confirms  
the programmed bolus amount of the “Quick”  
Standard bolus  
(Long vibration)  
You reset the “Quick” Standard Bolus amount to  
zero, you cancel the “Quick” Standard Bolus, or  
when the STOP-Warning occurs  
See the sections “Programming a Standard Bolus” (page 77) and “Setting Up  
Alarm Signals” (page 117) for more information.  
Appendixes  
182  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix F: Symbols  
General Symbols  
Read user guide  
Sterilized using ethylene oxide  
Sterilized using radiation  
Sterilized using antiseptics  
Year of manufacture  
Batch number  
Expiration date  
Item number  
Serial number  
Admissible temperature range during operation  
Admissible temperature range  
Admissible humidity range  
Admissible air pressure range  
Fragile – handle with care  
Protect against moisture  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
183  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Protect from heat and sunlight  
For single use only  
Does not contain pyrogen  
Does not contain PVC  
Recycling  
Do not throw away  
Do not use if the package is damaged  
Flammable  
See  
Electronic device of type BF according to the standard  
IEC 60601-1. Protection against electrical shock.  
Symbol for protection against the effect of temporary immersion  
in water [up to 60 minutes and 2.5 meters (8 feet)], according  
to IEC 60529.  
Marking of conformity according to the European Medical  
Device Directive MDD 93/42/EEC with the number of the  
notified body.  
Federal law (USA) restricts this device to sale by or on the  
order of a physician.  
Federal Communications Commission, part 15 on radio  
frequency devices.  
Manufacturer  
Appendixes  
184  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Symbol Definition  
Start your ACCU-CHEK Spirit insulin pump  
STOP screen and Stop insulin delivery  
Time, and Set the time and date menu  
Date  
Locked keys when the KeyLock function is turned on  
Unlocked keys when the KeyLock function is turned on  
Change the cartridge menu and cartridge content  
Cartridge low warning  
Empty cartridge  
Low battery  
Empty battery  
Alkaline AA battery  
Rechargeable AA battery  
Amount of International Units of insulin delivered per hour  
Basal rate profile  
Temporary Basal Rate  
TBR percentage  
Daily basal rate total  
Basal rate programming unlocked  
Basal rate programming locked  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
185  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Standard Bolus and Bolus increment programming  
Extended Bolus and Extended delivery of the MultiWave Bolus  
MultiWave Bolus  
Immediate bolus delivery of MultiWave Bolus  
Decreased Temporary Basal Rate (0–90%)  
Increased Temporary Basal Rate (110–200%)  
Remaining duration of an ongoing Extended Bolus, MultiWave  
Bolus, TBR, or automatic off  
Daily total of insulin delivered (basal rate plus boluses)  
Prime the infusion set (menu)  
Alert or error occurring  
Alert  
Error  
Information menu  
Pump timer  
Select the user menus STANDARD, ADVANCED or CUSTOM  
SETUP MENU STANDARD menu  
SETUP MENU ADVANCED menu  
Beeps turned on  
Vibrations turned on  
Beeps and vibrations turned on  
Appendixes  
186  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Set beep volume  
Screen orientation  
European date format  
American date format  
Alarm clock turned on  
Alarm clock turned off  
Single alarm when setting an alarm clock  
Every day repeat alarm when setting an alarm clock  
PC communication  
Set beep volume  
Check key  
Up key  
Down key  
Timeout of a menu  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
187  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix G: Sterile Products and Accessories  
Sterile Products  
Cartridges  
Name  
Comments  
Empty ACCU-CHEK  
3.15ml Plastic Cartridge  
Plastic cartridges are single-use  
items. Do not reuse cartridges.  
Refer to the instructions of the  
insulin you are using for information  
of proper use and acceptable  
temperature range for storage  
and handling.  
ACCU-CHEK infusion sets  
Name  
Comments  
ACCU-CHEK Ultraflex  
The ACCU-CHEK Ultraflex infusion  
set has a specially tapered cannula  
and reversible connector for easy,  
comfortable insertion and wearing  
convenience.  
ACCU-CHEK Tender  
ACCU-CHEK Rapid-D  
Variable insertion angle and choice  
of cannula for maximum control  
and comfort.  
Ultra-fine stainless needle inserts  
easily with one hand and is a great  
option for new pumpers or those  
with vision or dexterity issues.  
Appendixes  
188  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Accessories  
Name  
Comments  
Adapter  
Replace your adapter with every  
10th cartridge.  
Battery  
If used in a typical usage pattern  
[50 U/day using U100 insulin;  
operating temp. of 72°F 6°F  
(22°C 3°C)] the battery life is  
approx. four weeks for alkaline.  
One 1.5 AA alkaline battery with a  
minimum capacity of 2500mAh  
Rechargeable Battery  
One AA NiMH rechargeable battery with  
a minimum capacity of 1500 mAh  
If used in a typical usage pattern  
[50 U/day using U100 insulin;  
operating temp. of 72°F 6°F  
(22°C 3°C)] the battery life is  
approx. one week for rechargeable  
batteries.  
Battery Cover  
Replace your battery cover with  
every 4th battery.  
Carrying Systems  
Disetronic Medical Systems offers  
a wide range of carrying system  
solutions to best fit your lifestyle.  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
189  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix H: Glossary  
Adapter  
The adapter connects the cartridge and infusion set. It has two seals for the  
cartridge compartment of your insulin pump. The two small vents on the  
adapter let air pressure equalize.  
Basal rate  
The amount of insulin needed to cover your own basic insulin needs. In insulin  
pump therapy, your basal rate is determined by your doctor or healthcare team  
and can be adjusted to meet your body’s specific needs throughout the day.  
Your basal rate is delivered by your insulin pump according to your personal  
basal rate profile or profiles.  
Basal rate profile  
Your pump can be set to deliver up to five different basal rate profiles to match  
changing insulin needs (for example, during the week and on the weekend). A  
basal rate profile consists of 24 programmed hourly basal rates.  
Basal rate total  
The sum of all 24 hourly basal rates in one basal rate profile is called the (daily)  
basal rate total.  
Bolus  
The amount of insulin delivered (in addition to the basal rate) to cover the  
intake of food and correct high blood glucose levels. The bolus amount is  
determined by your doctor or healthcare team’s guidelines, your blood glucose  
level, your food intake, and your activity level.  
Carrying system  
A variety of special carrying systems made of different materials are designed  
to suit your individual needs while wearing your pump.  
Cartridge  
The insulin reservoir of your pump. It holds 3.15 ml of rapid-acting insulin analog.  
When using U100 insulin, 3.15 ml is equal to 315 International Units of insulin.  
Cartridge compartment  
The opening in your pump for the cartridge.  
Appendixes  
190  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Daily insulin total  
The total amount of insulin delivered (basal rate plus boluses) in a 24-hour day,  
beginning at midnight. This amount does not include any insulin needed for the  
priming of infusion sets.  
Diabetes Data Management (DDM)  
Diabetes Data Management is the recording of the therapy-relevant data  
gained from your insulin delivery system and your blood glucose measurement  
system (such as an ACCU-CHEK blood glucose monitor) in order to analyze and  
illustrate this data on a PC or other communication device.  
Factory settings  
The standard configuration (“factory settings”) on a new pump. These settings  
can be adapted to your individual needs directly on your pump.  
Filling aid  
An accessory that can be used to help you fill an empty ACCU-CHEK 3.15 ml  
plastic cartridge from your vial of insulin.  
Free flow  
Free flow of insulin from the infusion set may occur when two conditions are  
met: first, if the cartridge with a connected infusion set and the piston rod of  
your pump are not correctly coupled and second, if you position your pump at  
a higher level than the infusion site.  
Headset (also known as cannula housing)  
The part of an infusion set that includes the soft cannula or needle. The  
infusion set tubing connects either directly to the headset, or connects to a  
shorter length of tubing permanently attached to the headset.  
Hourly basal rate  
The amount of insulin delivered by your pump in 3-minute intervals for a  
given hour.  
Infusion set  
Connects your pump to your body. The insulin is delivered from the cartridge  
through the infusion set tubing and cannula or needle into your subcutaneous  
tissue.  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
191  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Infusion site  
The place where the infusion set cannula or needle is inserted into your  
subcutaneous tissue to deliver insulin.  
Insulin  
A hormone that helps cells turn glucose into energy. Insulin is produced in the  
beta cells of the pancreas (also called islets of Langerhans).  
Insulin, rapid-acting (insulin analog)  
A type of insulin created using recombinant DNA technology. The onset time  
of insulin analog is 5 to 15 minutes.  
IPX8 according to IEC 60529  
Protection against the effects of temporary immersion in water. Ingress of water  
in quantities causing harmful effects is not possible when the device is temporarily  
immersed in water under standardized conditions [maximum of 60 minutes a  
day and 8 feet (2.5 meters) depth].  
Luer-lock connector  
A standardized fitting at the end of the infusion set, the cartridge, and the adapter  
that lets them be locked together without leaking.  
Personal settings  
Your pump must be programmed with your personal settings prior to starting  
insulin pump therapy. Personal settings include the basal rate profile(s), the  
correct time and date, and all other variable values that can be tailored to  
your individual needs on your pump.  
RUN mode  
During normal use, your pump is in RUN mode and insulin is continuously  
delivered. Boluses, Temporary Basal Rates, and almost all functions can be  
programmed from RUN mode.  
RUN screen  
The RUN screen is the starting point for all functions that can be accessed  
while your pump delivers insulin. Your pump displays the RUN screen during  
normal use when no programming is currently being performed. The time,  
current hourly basal rate, basal rate profile, and all functions currently  
activated appear on the RUN screen.  
Appendixes  
192  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings  
Settings are individually programmable values and parameters that affect the  
way your pump works.  
Scrolling  
Scrolling allows fast and easy programming of larger values.  
STOP mode  
When your pump is in STOP mode, it does not deliver any insulin. The insulin  
delivery is only stopped if an error occurs or if your pump is put into STOP  
mode (for example, when changing the cartridge, adapter or infusion set, or  
for data transfer). Functions such as Extended Bolus or Temporary Basal Rate  
are interrupted by putting your pump into STOP mode.  
STOP screen  
The STOP screen is the starting point for all functions that require that your  
pump does not deliver insulin.  
Temporary Basal Rate  
Temporary increase or decrease of your basal rate profile in percentages (from  
0 – 200%) to match changing insulin needs due to increased or decreased  
activity level, illness, or stress.  
Timeout  
For your safety and convenience, your pump returns to the RUN or STOP  
screen if no key is pressed within a certain time. Any changes made are not  
saved.  
U100  
Each milliliter of liquid contains 100 International Units of insulin. Your pump  
was developed only to use U100 rapid-acting insulin analogs.  
Appendixes  
ACCU-CHEK Spirit Insulin Pump  
193  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LIMITED FOUR-YEAR WARRANTY  
DISETRONIC MEDICAL SYSTEMS, INC. (“DISETRONIC”) WARRANTS THE ACCU-CHEK SPIRIT  
INSULIN PUMP (“INSULIN PUMP”) AGAINST DEFECTS IN MATERIALS AND WORKMANSHIP FOR A  
PERIOD OF FOUR (4) YEARS FROM THE DATE OF PURCHASE. PROOF OF PURCHASE DATE MAY  
BE REQUIRED. THIS WARRANTY DOES NOT INCLUDE SUPPLIES AND ACCESSORIES, INCLUDING  
BUT NOT LIMITED TO, CARTRIDGES, BATTERIES, OR INFUSION SETS.  
DURING THE WARRANTY PERIOD, DISETRONIC WILL REPAIR OR REPLACE, IN ITS SOLE  
DISCRETION, ANY INSULIN PUMP THAT DOES NOT WORK PROPERLY BECAUSE OF A DEFECT IN  
MATERIALS OR WORKMANSHIP. THE PURCHASER’S SOLE AND EXCLUSIVE REMEDY WITH  
RESPECT TO THE INSULIN PUMP SHALL BE REPAIR OR REPLACEMENT. THIS WARRANTY APPLIES  
ONLY TO NEW DEVICES. IN THE EVENT A PUMP IS REPAIRED OR REPLACED, THE WARRANTY  
PERIOD WILL NOT BE EXTENDED.  
THIS WARRANTY IS VALID ONLY IF THE INSULIN PUMP IS USED IN ACCORDANCE WITH THE USER  
INSTRUCTIONS. THIS WARRANTY WILL NOT APPLY:  
1. IF DAMAGE RESULTS FROM CHANGES OR MODIFICATIONS MADE TO THE INSULIN PUMP BY  
THE USER OR THIRD PERSONS AFTER THE DATE OF MANUFACTURE;  
2. IF DAMAGE RESULTS FROM SERVICE OR REPAIRS PERFORMED BY ANY PERSON OR ENTITY  
OTHER THAN DISETRONIC.  
3. IF DAMAGE RESULTS FROM A FORCE MAJEURE OR OTHER EVENT BEYOND THE CONTROL OF  
DISETRONIC; OR  
4. IF DAMAGE RESULTS FROM NEGLIGENCE OR IMPROPER USE, INCLUDING BUT NOT LIMITED  
TO IMPROPER STORAGE, PHYSICAL ABUSE SUCH AS DROPPING, OR OTHERWISE.  
THIS WARRANTY SHALL BE PERSONAL TO THE ORIGINAL PURCHASER. ANY SALE, RENTAL, OR  
OTHER TRANSFER OR USE OF THE INSULIN PUMP COVERED BY THIS WARRANTY TO OR BY A  
USER OTHER THAN THE ORIGINAL PURCHASER SHALL CAUSE THIS WARRANTY TO IMMEDIATELY  
TERMINATE.  
THIS WARRANTY IS EXCLUSIVE OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED INCLUDING  
WITHOUT LIMITATION THE IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A  
PARTICULAR PURPOSE. IN NO EVENT SHALL DISETRONIC OR ITS SUPPLIERS OR DISTRIBUTORS  
BE LIABLE FOR ANY INCIDENTAL, CONSEQUENTIAL, INDIRECT, SPECIAL OR PUNITIVE DAMAGES  
ARISING FROM OR IN ANY WAY CONNECTED WITH THE PURCHASE OR USE OF THE INSULIN  
PUMP.  
Appendixes  
194  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Cartridge; 9, 18  
Changing; 73-74  
Index  
Connecting; 47-48  
Filling; 43-46  
Inserting; 49-52  
Low; 149, 185  
Accessories; 17-23, 189  
Adapter; 9, 20, 47-48, 190  
Advanced User Menu; 99, 122-128  
Air Bubbles; 45, 53, 145  
Alarm Clock; 127-128, 151  
Alarm History; 109  
CHECK Key; 9, 187  
Cleaning; 141  
Copying Hourly Basal Rate; 40-41  
Alarm Signals; 16, 117  
Alcohol Consumption; 144  
Alerts and Errors; 147-164, 199-200  
Automatic Off; 118, 156  
Data Interrupted; 163  
Data Transfer; 104  
Date; 32-34, 150  
Date Format; 130  
Daily Insulin Totals; 110, 191  
Display; 9-10  
Contrast; 135  
Orientation; 120, 187  
Disposal; 140  
Backlight; 10  
Basal Rate Profile; 36, 123, 126, 190  
Lock; 133  
Programming; 37-41, 124  
Selecting; 125  
Temporary; 93-96  
Battery; 21, 141  
DOWN Key; 9, 187  
Cover; 22  
Depleted; 155  
Inserting or changing; 25-27  
Key; 23, 71  
Life; 165, 189  
Electromagnetic; 61, 169-172  
Electronic Error; 160  
Emergency Kit; 23  
Errors; 147-148, 154-164  
Exit Options; 102  
Low; 150, 185  
Type; 119, 167  
Beeps; 15, 116, 181, 186-187  
Bolus; 75  
Filling Aid; 43  
Guarantee; 4  
Extended; 83-87  
History; 107-108  
MultiWave; 88-92  
“Quick” Standard; 77-80  
“Scroll” Standard; 80-82  
History; 107  
Alarm; 109  
Bolus; 107-108  
Daily Insulin; 110  
Temporary Basal Rate; 111  
Index  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
195  
Immunity; 171-172  
Infusion Set; 9, 19-20  
Power Interrupt; 161  
Prime Quantity; 132  
Priming; 53-55  
Changing the; 70-72  
Not Primed; 162-163  
Priming; 53-55  
Infusion Site; 56-57, 72  
Insulin Delivery;  
Starting; 65  
Products, Sterile; 17-18, 188  
Profile, See Basal Rate Profile  
Programming;  
Additional Basal Rate  
Profile; 124  
Stopping; 66  
Insulin Pump Therapy;  
Basal Rate Profile; 37  
Extended Bolus; 83-87  
MultiWave Bolus; 88-92  
“Quick” Standard Bolus;  
77-80  
Interruption of; 68-69  
Resuming; 69  
KeyLock; 13  
“Scroll” Standard Bolus;  
80-82  
Temporary Basal Rate; 93-96  
Pump; 9  
Turning On; 113-114  
Turning Off; 113-114  
Unlocking; 181, 185  
Keys; 11  
Stopping; 66  
Key Combinations; 12  
Storing; 140  
Wearing; 59-60  
Language; 134  
Language Error; 164  
Liquids; 64  
Quick Info Screen; 14  
Quick Information; 30  
Luer-Lock; 192  
Repair; 142  
Maintenance; 141  
Mechanical Error; 159-160  
Menus; 99-102  
RUN Screen; 13, 192  
Safety System; 147, 168  
Screen; 9-10  
Advanced; 122  
Custom; 136  
Scrolling; 193  
Standard; 103  
Setup Menu;  
MultiWave Bolus;  
Programming; 88-91  
Cancelling; 91-92  
Advanced; 129, 186  
Standard; 113, 186  
Standard User Menu; 99, 103  
Start-Up; 28-31  
Occlusion; 146  
Error; 156-158, 167  
Orientation; 120  
Status Screen; 15  
Sterile; 17, 188  
Stop-Warning; 16, 67  
Index  
196  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Stopping;  
Insulin Delivery; 66  
Quick Reference Table for Errors  
Priming; 55  
System Check; 28-31  
No. Error  
What You Need to Do  
E1  
E2  
E3  
CARTRIDGE EMPTY  
Change the cartridge.  
Change the battery.  
Temporary Basal Rate;  
Cancelling; 97  
History; 111  
BATTERY DEPLETED  
AUTOMATIC OFF  
If necessary, put your ACCU-CHEK Spirit in  
RUN mode.  
Over; 96, 153  
Programming; 93-96  
Time and Date; 32  
American; 32, 130  
European; 32, 130  
Time and Date, Review; 150  
E4  
E5  
OCCLUSION  
Immediately check your blood glucose level,  
because insulin delivery has been interrupted.  
If your blood glucose level is high, take  
appropriate actions according to your doctor’s  
or healthcare team’s instructions. See the  
section “Error E4: OCCLUSION.”  
Time Remaining; 112  
Troubleshooting; 143-146  
Tubing; 20, 132  
END OF OPERATION  
Arrange for immediate replacement of your  
insulin pump. Contact Pump  
Support at  
1-800-688-4578 to dis-  
cuss your options for further use of the  
ACCU-CHEK Spirit insulin pump. Contact  
your healthcare professional for alternative  
insulin therapy options.  
UP Key; 9, 187  
User Menu; 99-102  
Advanced; 122  
Custom; 136  
E6  
MECHANICAL ERROR  
ELECTRONIC ERROR  
POWER INTERRUPT  
CARTRIDGE ERROR  
See the section “Error E6: MECHANICAL  
ERROR.”  
Standard; 103  
E7  
See the section “Error E7: ELECTRONIC  
ERROR.”  
Water; 63  
Contact with; 63  
After Contact with; 64  
E8  
See the section “Error E8: POWER INTER-  
RUPT.”  
E10  
See the section “Error E10: CARTRIDGE  
ERROR.”  
E11  
E12  
E13  
SET NOT PRIMED  
DATA INTERRUPTED  
LANGUAGE ERROR  
Prime your infusion set.  
Restart the data transfer.  
See the section “Error E13: LANGUAGE  
ERROR.”  
Index  
ACCU-CHEK Spirit Insulin Pump  
197  
Quick Reference Table for Errors  
200  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Quick Reference Table for Alerts  
Diagram of User Menus  
In all cases, turn off and confirm the alert prior to addressing the issue.  
Your Insulin pump STANDARD MENU  
No. Alert  
What You Need to Do  
KeyLock  
A1  
CARTRIDGE LOW  
Change cartridge before it is completely  
empty.  
quick info  
Standard  
bolus  
KeyLock  
quick info  
«Scroll» Stan-  
dard bolus  
temporary  
basal rate  
stop your  
information  
menu  
program  
set the time  
and date  
setup menu  
standard  
select the  
user menu  
A2  
A3  
A4  
BATTERY LOW  
Replace the battery as soon as possible.  
change the  
cartridge  
prime the  
communi-  
cation  
Accu-Chek  
profile no.1  
infusion set  
Standard  
bolus  
REVIEW TIME AND DATE Set the time and date.  
bolus  
amount  
percentage  
duration  
bolus  
history  
daily basal  
rate total  
set the  
hour  
KeyLock  
user menu  
start your  
remove the  
cartridge  
start  
priming  
data transfer  
to PC  
Accu-Chek  
ALARM CLOCK  
Put your ACCU-CHEK Spirit in RUN mode,  
if necessary.  
alarm  
history  
1st hour  
2nd hour  
set the  
minute  
beep  
volume  
BR profile  
no.…  
returning the  
piston rod  
stop  
priming  
(A) «Quick»  
with direct access from the RUN screen  
Standard bolus using the a and s keys of your  
insulin pump.  
daily insulin  
totals  
set the  
year  
alarm  
signals  
fill volume  
315 units  
A5  
A6  
PUMP TIMER  
Your pump’s operating time will expire  
soon. Go to the INFORMATION screen on  
your pump, then move to the PUMP TIMER  
screen. The days until the pump timer  
expires is shown. Make arrangements for  
replacement of the pump before the timer  
reaches zero days.  
(B) «Scroll»  
menu-guided using the STANDARD BOLUS  
TBR history  
set the  
month  
automatic  
off  
start-up  
procedure  
Standard bolus menu and the scroll function of the a and s  
keys for programming the bolus amount.  
maintenance  
24th hour  
set the  
day  
battery  
type  
For a complete review of your insulin pump features as well as  
the warnings and precautions associated with its use, please  
refer to your ACCU-CHEK Spirit Insulin Pump User Guide.  
display  
orientation  
Your Insulin pump ADVANCED MENU  
KeyLock  
TBR CANCELLED  
Put your ACCU-CHEK Spirit insulin pump in  
RUN mode, if necessary. Ensure the can-  
cellation was intended and program a new  
Temporary Basal Rate, if necessary.  
quick info  
Standard  
bolus  
KeyLock  
quick info  
«Scroll» Stan-  
dard bolus  
Extended  
bolus  
MultiWave  
bolus  
temporary  
basal rate  
stop your  
Accu-Chek  
information  
menu  
change your  
BR profile  
program
profileno.12345  
alarm clock  
set the time  
and date  
setup menu  
standard  
setup menu  
advanced  
select the  
user menu  
change the  
cartridge  
prime the  
infusion set  
communi-  
cation  
bolus  
amount  
bolus  
amount  
bolus  
amount  
percentage  
duration  
bolus  
history  
BR profile  
no. …  
daily basal  
rate total  
alarm  
occurance  
set the  
hour  
KeyLock  
time  
format  
user menu  
start your  
Accu-Chek  
remove the  
cartridge  
start  
priming  
data transfer  
to PC  
A7  
A8  
TBR OVER  
Decide whether a further Temporary Basal  
Rate change is appropriate and program  
one, if necessary.  
bolus  
duration  
immediate  
bolus  
alarm  
history  
1st hour  
2nd hour  
alarm  
hour  
set the  
minute  
beep  
volume  
date  
format  
BR profile  
no.…  
returning the  
piston rod  
stop  
priming  
bolus  
duration  
fill volume  
315 units  
daily insulin  
totals  
alarm  
minute  
set the  
year  
alarm  
signals  
bolus  
increment  
BOLUS CANCELLED  
Put your ACCU-CHEK Spirit insulin pump in  
RUN mode, if necessary. Ensure that the  
cancellation was intended and program a  
new bolus, if necessary.  
TBR history  
set the  
month  
automatic  
off  
prime  
quantity  
start-up  
procedure  
maintenance  
24th hour  
set the  
day  
battery  
type  
basal  
rate lock  
display  
orientation  
language  
display  
contrast  
Quick Reference Table for Alerts  
ACCU-CHEK Spirit Insulin Pump  
199  
Index  
198  
ACCU-CHEK Spirit Insulin Pump  
Download from Www.Somanuals.com. All Manuals Search And Download.  

321 Studios Coffeemaker 4DX 33 User Manual
AeroGarden Greenhouse Kit 100610 BLK User Manual
AKG Acoustics Microphone D7 User Manual
Allied Air Enterprises Air Conditioner 4AC16LT User Manual
Amcor Air Conditioner Single Zone Inverter Systems User Manual
AMD Computer Hardware PCM 5820 User Manual
American Standard Plumbing Product 7415801 User Manual
Ativa Paper Shredder DSD160D User Manual
Atlas Sound Switch AM1100 User Manual
Axxess Interface Automobile Accessories SM5 TY6009G3 User Manual