		 Pump User Guide   
					ACCU-CHEK™ Spirit   
					insulin pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Dear ACCU-CHEK™ Spirit insulin pump owner,   
					Thank you for choosing the ACCU-CHEK Spirit insulin pump.   
					This guide will help you to get the most from your ACCU-CHEK Spirit insulin   
					pump. Even if you have used an insulin pump before, it’s important that you   
					do these things before you use the ACCU-CHEK Spirit insulin pump:   
					̈ read this guide carefully   
					̈ work with your doctor or healthcare team to understand how to use   
					the pump   
					̈ test your blood glucose level often to find your personal settings and   
					to avoid getting the wrong dosage of insulin. (An ACCU-CHEK blood   
					glucose meter is one way to test blood glucose levels; talk to your   
					doctor or healthcare team to decide the best way for you.)   
					It’s a good idea to inform family members or friends about diabetes and pump   
					therapy so that they can help you in an emergency.   
					Disetronic Medical Systems is proud to help you manage your diabetes therapy.   
					However, Disetronic Medical Systems cannot recommend specific diabetes   
					therapy. Always follow the instructions given to you by your doctor or healthcare   
					team. If you have questions about your pump, or problems with it, refer to this   
					guide and talk to your doctor or healthcare team. We also invite you to contact   
					Pump Support at 1-800-688-4578.   
					Note:   
					Your insulin pump is a valuable medical device. Consider adding your   
					pump to your homeowner’s insurance policy to protect you in case it is   
					damaged or lost. Ask your insurance agent for details.   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				How to use this guide   
					This guide is for version 2.XX of the ACCU-CHEK Spirit insulin pump.   
					Note:   
					This user guide presents you with sample screens. The screens in this   
					user guide may look slightly different from your pump screens. If you   
					have questions about what you see on your pump, contact Pump   
					Support at 1-800-688-4578.   
					When you become familiar with your pump, you can quickly refer to the back   
					of this user guide for tables and definitions that can help you to understand   
					pump alerts and warnings, and the symbols on the pump screen. There is also   
					a glossary at the back that defines terms used in this guide, and a table that   
					lists the technical specifications for your pump. All of this information can help   
					you and your doctor or healthcare team to understand the pump and use it to   
					best manage your insulin pump therapy.   
					Warnings and Cautions   
					Read all warnings and cautions in this guide before you use the pump. This   
					information is important to protect your health and the function of your pump.   
					If you have any questions about a warning or caution, contact Pump Support   
					at 1-800-688-4578. Warnings and Cautions are indicated by triangles.   
					WARNINGS are indicated by a triangle and the word   
					WARNING.   
					! 
					L 
					WARNING   
					CAUTIONS are indicated by a triangle and the word   
					CAUTION.   
					! 
					L 
					CAUTION   
					2 
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Working safely with your ACCU-CHEK Spirit insulin pump   
					Your ACCU-CHEK Spirit insulin pump meets the safety requirements of national   
					medical device legislation and meets or exceeds the international standards for   
					Electromagnetic Compatibility regarding the pump’s use. This section introduces   
					you to the intended use of the ACCU-CHEK Spirit insulin pump, to its safety   
					features, and to warnings and safety information you need to know while using   
					your pump.   
					While your pump is in use, two internal safety systems monitor and control the   
					pump’s functioning. The pump performs more than nine million safety checks   
					every day. If your pump detects a change from its normal state, an alert or an   
					error message is delivered.   
					See the sections “Alerts and Errors” (page 147) and “Appendix A: Technical Data”   
					(page 165) for information on how to address errors and alerts.   
					Intended use of your ACCU-CHEK Spirit insulin pump   
					The ACCU-CHEK Spirit insulin pump is intended for the subcutaneous continuous   
					delivery of insulin, at set and variable rates, for the management of diabetes   
					mellitus in persons requiring insulin as prescribed by a physician.   
					The ACCU-CHEK Spirit insulin pump is intended for the use of U100 rapid-   
					acting insulin.   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					3 
				Guarantee   
					Any changes or modifications to the ACCU-CHEK Spirit insulin pump that   
					are not expressly approved by Disetronic Medical Systems, Inc. could void   
					the pump warranty. Do not try to fix the device yourself.   
					4 
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Contents   
					Chapter 1: Understanding Your New System _____________________9   
					The ACCU-CHEK Spirit Insulin Pump __________________________9   
					Display ________________________________________________9   
					
					Keys and Key Combinations _______________________________11   
					
					
					Alarm Signals __________________________________________16   
					Accessories and Disposables ______________________________17   
					Emergency Kit __________________________________________23   
					Chapter 2: Getting Started __________________________________25   
					Inserting (or Changing) the Battery__________________________25   
					System Check Start-Up Process ____________________________28   
					Setting Time and Date____________________________________32   
					Your Personal Settings ___________________________________35   
					Your Basal Rate Profile ___________________________________36   
					Programming Your Basal Rate Profile________________________37   
					Preparing the Cartridge___________________________________42   
					Connecting the Cartridge, Adapter, and Infusion Set_____________47   
					Inserting the Cartridge ___________________________________49   
					Priming the Infusion Set __________________________________53   
					Preparing the Infusion Site and Inserting the Infusion Set ________56   
					Chapter 3: Using Your Pump_________________________________59   
					Before Using ___________________________________________59   
					Wearing Your Pump______________________________________59   
					When Not to Wear Your Pump ______________________________61   
					Your Insulin Pump and Water ______________________________63   
					Starting Insulin Delivery __________________________________65   
					Stopping Insulin Delivery _________________________________66   
					Interruption of Insulin Pump Therapy ________________________68   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					5 
				Changing the Infusion Set_________________________________70   
					Changing the Cartridge and Infusion Set _____________________73   
					Chapter 4: Boluses and Temporary Basal Rates__________________75   
					Understanding Boluses ___________________________________75   
					Programming a Standard Bolus ____________________________77   
					Programming an Extended Bolus ___________________________83   
					Programming a MultiWave Bolus ___________________________88   
					Programming a Temporary Basal Rate (TBR) __________________93   
					Chapter 5: User Menus _____________________________________99   
					Selecting a User Menu __________________________________100   
					STANDARD User Menu___________________________________103   
					Data Transfer__________________________________________104   
					Setup for Data Transfer __________________________________105   
					Reviewing Pump Data ___________________________________107   
					SETUP MENU Standard __________________________________113   
					ADVANCED User Menu___________________________________122   
					Basal Rate Profiles _____________________________________123   
					Alarm Clock___________________________________________127   
					SETUP MENU Advanced__________________________________129   
					CUSTOM User Menu_____________________________________136   
					Chapter 6: Care and Maintenance ___________________________137   
					Care of Your ACCU-CHEK Spirit Insulin Pump _________________137   
					System Inspection Checklist______________________________137   
					What to Do When You Drop Your Pump ______________________139   
					Storing Your ACCU-CHEK Spirit Insulin Pump _________________140   
					Disposal _____________________________________________140   
					Maintenance __________________________________________141   
					Repair _______________________________________________142   
					Troubleshooting________________________________________143   
					6 
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Chapter 7: Alerts and Errors________________________________147   
					Confirming an Alert or Error ______________________________148   
					List of Alerts __________________________________________149   
					List of Errors __________________________________________154   
					Appendixes_____________________________________________165   
					Appendix A: Technical Data_______________________________165   
					Appendix B: Configuration Parameters ______________________177   
					Appendix C: Abbreviations _______________________________180   
					Appendix D: Beeps and Melodies __________________________181   
					Appendix E: Vibrations __________________________________182   
					Appendix F: Symbols____________________________________183   
					Appendix G: Sterile Products and Accessories ________________188   
					Appendix H: Glossary ___________________________________190   
					Warranty _____________________________________________194   
					Index _________________________________________________195   
					Diagram of User Menus ___________________________________198   
					Quick Reference Table for Alerts ____________________________199   
					Quick Reference Table for Errors ____________________________200   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					7 
				This page left blank intentionally.   
					8 
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Chapter 1: Understanding Your New System   
					The ACCU-CHEK Spirit Insulin Pump   
					Display   
					Menu key   
					Up key   
					Down key   
					Check key   
					Infusion set   
					Adapter   
					Cartridge   
					Check your insulin pump and its sterile products, like an   
					infusion set and accessories, at least every three hours.   
					! 
					L 
					Immediately inspect your insulin pump if it is dropped.   
					Do not use your insulin pump if chips or cracks are visible.   
					Chips and cracks may allow water, dust, insulin, or other   
					foreign substances to enter your insulin pump and lead   
					to malfunction.   
					CAUTION   
					Display   
					Your insulin pump has a Liquid Crystal Display (LCD) screen that displays   
					important information. Check the display at least every three hours during   
					the day, before you go to sleep, and if you are not able to hear the beeps or   
					alerts or feel the vibrations. This is the only way you will be aware of important   
					changes in your pump.   
					Chapter 1: Understanding Your New System   
					ACCU-CHEK Spirit Insulin Pump   
					9 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				If you discover incomplete characters, numbers, or symbols   
					on the display, put your insulin pump into STOP mode and   
					perform a system check by simply removing the battery for a   
					few seconds. When you reinsert the battery the system check   
					will occur. If the incomplete characters, numbers, or symbols   
					remain, your insulin pump may need to be returned since you   
					may not receive the correct information for operating your   
					pump. Put your insulin pump into STOP mode and contact   
					Pump Support at 1-800-688-4578 immediately.   
					! 
					L 
					CAUTION   
					Note:   
					To make the display easier to read, you can flip the display orientation   
					by 180° and adjust the display contrast lighter or darker. See the sections   
					“Display Orientation” (page 120) and “Display Contrast” (page 135) for   
					more information.   
					Backlight   
					The backlight helps you to see the display in poorly lighted areas. You can turn   
					on the backlight from the RUN screen, from the STOP screen, or when you are   
					browsing the menus. Press the a key to turn on the backlight. The backlight   
					turns itself on when an alert or an error message appears. The backlight turns   
					itself off after 30 seconds if no keys are pressed.   
					Hint:   
					Start programming with the a key in areas of poor lighting as this will   
					also turn on the backlight.   
					Chapter 1: Understanding Your New System   
					10   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Keys and Key Combinations   
					The pump will beep each time a key is pressed, unless the beep tone is set to 0.   
					Never press any of the keys with a sharp or pointed item   
					such as the tip of your fingernail. This could damage the   
					casing of your insulin pump. Use the pad of your fingertip   
					instead to avoid puncturing the keys.   
					! 
					L 
					CAUTION   
					Single Keys   
					Each key has a specific function when it is pressed by itself. The following   
					table describes the function of each.   
					Key Name Function   
					d 
					Menu   
					scrolling through menus, functions, and   
					information screens   
					f 
					Check   
					• selecting a menu   
					• saving changes and exiting screen   
					• viewing the QUICK INFO screen   
					a 
					s 
					Up   
					• increasing a setting   
					• moving forward in the information screens   
					• turning on the backlight   
					Down   
					• decreasing a setting   
					• moving backward in the information screens   
					Either a or s can be used for programming or cancelling a Standard   
					bolus. Pressing either key for 3 seconds (until you hear a melody) will   
					also temporarily turn off the STOP-Warning.   
					Chapter 1: Understanding Your New System   
					ACCU-CHEK Spirit Insulin Pump   
					11   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					 
				Note:   
					Increase or decrease the values by pressing the a or s keys. If   
					you want to make large changes to a value, in most cases you can do   
					this faster by pressing and holding down the a or s key (scrolling)   
					until the number you want is displayed. If needed, the value can still   
					be corrected by a single press of the s or a. Your pump will beep   
					once when you start scrolling.   
					Key Combinations   
					Keys can be pressed in pairs for additional functions. In the following table,   
					a plus symbol (+) between 2 keys means that both keys must be pressed   
					simultaneously.   
					Combination Function   
					d + a   
					d + s   
					a + s   
					• exiting menus, function screens, and   
					information screens without saving   
					• moving backwards to a previous menu   
					unlocking the keys (press and hold both   
					keys until you hear three beeps in RUN mode   
					and one beep in STOP mode.)   
					copying an hourly basal rate to the next hour(s)   
					If the keys on your pump are not functioning properly, or you   
					are not able to identify them correctly, disconnect your   
					insulin pump and contact Pump Support at 1-800-688-4578.   
					! 
					L 
					WARNING   
					Chapter 1: Understanding Your New System   
					12   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Locking Keys Using KeyLock Feature   
					The KeyLock feature lets you lock all four keys on your pump as a safety measure   
					against accidentally activating functions. You may want to use this function   
					while sleeping, or during sports, for example. The KeyLock function is turned on   
					in the STANDARD SETUP MENU. All instructions in this user guide are given on   
					the assumption that the KeyLock is off. For more information see the section   
					“Turning KeyLock On or Off” (page 113).   
					Screens   
					Your pump displays various screens on its display for you to access information   
					and functions.   
					RUN Screen   
					basal rate   
					time   
					current hourly   
					basal rate   
					KeyLock locked   
					pump timer   
					cartridge low   
					battery low   
					active basal rate profile   
					During normal use, your pump is in RUN mode and insulin is delivered into   
					your body based on the basal rate you have programmed. The RUN screen is   
					the “home” screen and is displayed when the pump is in RUN mode. After you   
					program or review information, the pump returns to the RUN screen. This   
					screen shows the current time, the current basal rate profile, and the current   
					hourly basal rate in units per hour. The RUN screen will also display reminder   
					information (such as “cartridge low”) and special functions (such as KeyLock   
					locked or unlocked) in the form of symbols.   
					See the sections “Starting Insulin Delivery” (page 65) for instructions on starting   
					your pump and “Appendix F: Symbols” (page 183) for more information about   
					the symbols.   
					Chapter 1: Understanding Your New System   
					ACCU-CHEK Spirit Insulin Pump   
					13   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				STOP Screen   
					date   
					time   
					KeyLock unlocked   
					pump timer   
					cartridge low   
					battery low   
					The STOP screen shows the current time and date, the STOP symbol, reminder   
					information (such as “battery low”), and special functions (such as KeyLock locked   
					or unlocked). Some functions must be programmed while your pump is in STOP   
					mode, when insulin is not being delivered. Your pump must also be in STOP mode   
					before you can change cartridges. Additionally, always disconnect your pump from   
					the infusion site and place the pump in STOP mode when you prime the infusion   
					set and/or when you transfer data. See the section “Stopping Insulin Delivery”   
					(page 66) for instructions on stopping your pump.   
					QUICK INFO Screen   
					If you press the f key from the RUN or STOP   
					screen, your pump displays the remaining cartridge   
					content in units. Symbols for activated functions and   
					features (like the beep and vibration alarm signals) are displayed. Your pump will   
					return to the RUN or STOP screen after 8 seconds.   
					Information, Function, and Setup Screens   
					You will need to access various screens so you can view stored information,   
					program the pump, perform a function, or change a setting. You can reach   
					these screens from either the RUN or STOP screen.   
					Chapter 1: Understanding Your New System   
					14   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Follow the steps below to navigate through the screens.   
					1. Press the d key to move to a screen displaying your first choice. Think   
					of these screens as menu choices.   
					2. Press the d key again to see your next choice on the menu. After you   
					reach the last menu choice you will “loop” back to the first menu choice.   
					3. When you reach the screen displaying the choice you want (e.g., SET THE   
					TIME AND DATE), press f. The next set of screens are called function   
					screens. They all relate to the choice you made. There may be only one   
					function screen or multiple screens.   
					4. Press the a or s key to increase or decrease the selected value. Or,   
					press the d key again to move to the next function screen you want to   
					change. After you reach the last function screen you will “loop” back to   
					the first function screen.   
					5. Press f to save your changes. The pump returns to the RUN or STOP   
					screen and the changes are saved.   
					Note:   
					The menu choices that appear depend on the menu that is selected.   
					Your pump can display a Standard, Advance, or Custom user menu.   
					See the section “User Menus” (page 99) for more information.   
					Pressing d+a simultaneously lets you move backwards in the   
					menu structure or to return to menus you just skipped. If you are   
					within a function screen, pressing d+a simultaneously lets you   
					move up to the corresponding menu without saving the current changes.   
					STATUS Screen Messages, Beeps, and Vibrations   
					Beeps, vibrations, and messages on the display inform you of the status of   
					your insulin pump. Beeps and/or vibrations are used as alarm signals. Your   
					insulin pump also beeps when a key is pressed unless the beep volume is   
					turned off. The beep volume can be turned up or down.   
					Chapter 1: Understanding Your New System   
					ACCU-CHEK Spirit Insulin Pump   
					15   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Check the display of your insulin pump at least every three   
					hours during the day and before you go to sleep, especially   
					if for any reason you are unable to hear the beeps or feel   
					the vibrations. This is the only way you will be notified of   
					changes in your insulin pump in a suitable amount of time.   
					! 
					L 
					WARNING   
					Alarm Signals   
					Your pump gives alerts and errors using beeps and vibrations. You can turn off   
					either the beep or the vibrations but for your safety you cannot turn off both at   
					the same time for the initial alarm signal. See the sections “Alerts and Errors”   
					(page 147), “Adjusting the Beep Volume” (page 116), and “Setting Up Alarm   
					Signals” (page 117) for more information.   
					STOP-Warning   
					The STOP-Warning is turned on by the pump each time your pump switches from   
					RUN mode to STOP mode, and when a battery is inserted. The STOP-Warning   
					is a long beep and a vibration occurring every minute. This warns you that   
					insulin delivery has been interrupted.   
					To Turn Off the STOP-Warning   
					Press either a or s for 3 seconds to turn off the STOP-Warning. A melody   
					plays indicating the STOP-Warning is turned off.   
					Note:   
					The volume of the STOP-Warning cannot be changed. It is independent   
					of the programmed beep volume.   
					Chapter 1: Understanding Your New System   
					16   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Accessories and Disposables   
					Sterile products and accessories are a crucial part of the ACCU-CHEK Spirit   
					insulin pump system and therapy. These products are designed for safe and   
					simple insulin pump therapy. Always follow the instructions given by your   
					doctor or healthcare team for the use of all system components.   
					Use your ACCU-CHEK Spirit insulin pump, sterile products, and accessories   
					for the first time in the presence of your doctor or healthcare team.   
					Sterile Products   
					Each sterile product should be used only once. Reusing these products could   
					lead to infection or could damage your pump. Do not use a sterile product if   
					the package is damaged or opened.   
					Sterility is guaranteed up to the indicated expiration date for all ACCU-CHEK   
					sterile products that are in the original, unopened packaging.   
					Your insulin pump was designed for use with U100 insulin.   
					Only use sterile products and accessories that are designed   
					! 
					L 
					for use with your insulin pump. The proper functionality   
					of your insulin pump can only be guaranteed if you use   
					ACCU-CHEK brand sterile products and ACCU-CHEK Spirit   
					accessories. All ACCU-CHEK brand sterile products and   
					accessories designed for the ACCU-CHEK Spirit insulin pump   
					have been thoroughly tested and approved for use with   
					your insulin pump. Other sterile products and accessories   
					have not been tested for compatibility with your insulin   
					pump and, if used, may therefore endanger your health   
					or void your warranty.   
					WARNING   
					Chapter 1: Understanding Your New System   
					ACCU-CHEK Spirit Insulin Pump   
					17   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Always have extra sterile products and accessories with you.   
					This allows you to exchange components when needed.   
					Materials designed for single use (such as cartridges and   
					infusion sets) should not be reused due to the increased risk   
					of infections, malfunctions, and/or incorrect insulin delivery.   
					! 
					L 
					CAUTION   
					See Disetronic Medical Systems’ sterile products and accessories catalogs and   
					brochures for more information. Or, contact Pump Support at 1-800-688-4578.   
					Your insulin pump and its sterile products and accessories   
					include small component pieces that could pose a choking   
					hazard to children. Keep sterile products and accessories   
					out of the reach of children.   
					! 
					L 
					WARNING   
					Cartridges   
					Your pump is designed to use ACCU-CHEK 3.15 ml plastic cartridges to hold   
					insulin. Do not reuse a cartridge. Reusing sterile products could lead to infection   
					or could damage your pump.   
					ACCU-CHEK 3.15 ml plastic cartridges come with a protective cartridge cap,   
					as well as a filling needle and plunger rod to aid in filling.   
					cartridge   
					plunger rod   
					plunger   
					cartridge protective cap   
					filling needle with protective cap   
					Chapter 1: Understanding Your New System   
					18   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				The cartridges are sterile products intended for single   
					use only.   
					! 
					L 
					WARNING   
					Infusion Set   
					An ACCU-CHEK infusion set connects your pump to your body to deliver insulin.   
					The insulin travels from the cartridge in your insulin pump through the infusion   
					set into your body. The cannula or needle of the infusion set is placed under   
					the skin, usually in the abdominal area.   
					Recommended   
					Possible   
					Chapter 1: Understanding Your New System   
					ACCU-CHEK Spirit Insulin Pump   
					19   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Use only infusion sets that have a luer-lock connector. All current ACCU-CHEK   
					infusion sets have luer-lock connectors and are the preferred choice for use   
					with your ACCU-CHEK Spirit insulin pump. All ACCU-CHEK infusion sets are PVC   
					free, are skin-friendly, and do not affect insulin. The diameter of the ACCU-CHEK   
					infusion set tubing is small for fast priming and minimal waste of insulin.   
					ACCU-CHEK infusion sets come in a wide range of lengths from 60 cm (24 inches)   
					to 110 cm (43 inches). Choose your tubing length based on how you wear   
					your pump. Please refer to the infusion set instructions for proper usage and   
					priming volumes.   
					Adapter   
					Vents   
					Seal inside the adapter   
					Seal outside the adapter   
					The adapter connects the cartridge to the infusion set. It has two seals to keep   
					water from getting into the cartridge compartment. Two vents on the adapter   
					allow air pressure to equalize. The adapter will not work properly if the vents are   
					plugged, dirty, or if the seals are worn or missing.   
					Change the adapter with at least every 10th cartridge change. Also change the   
					adapter if the seals become worn, lost, or if the vents become plugged.   
					Chapter 1: Understanding Your New System   
					20   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Battery   
					Your pump runs on one AA battery. Use AA alkaline batteries that have a minimum   
					capacity of 2500 mAh. Do not use lithium, carbon zinc, or nickel cadmium (NiCd)   
					batteries. If you prefer to use rechargeable batteries, use NiMH batteries that   
					have a minimum capacity of 1500 mAh. If you use rechargeable batteries, always   
					use a battery charger recommended by the battery manufacturer. Keep extra   
					batteries on hand.   
					Using batteries other than those recommended by Disetronic   
					Medical Systems may significantly reduce battery usage   
					time. Batteries other than those recommended could leak   
					and corrode the battery contacts within your insulin pump.   
					Therefore, using batteries not supplied or recommended   
					by Disetronic Medical Systems, Inc. may void the warranty.   
					! 
					L 
					CAUTION   
					The operating temperature of the battery must be between 41°F and 104°F   
					(+5°C and +40°C). The life of any battery is affected by how you use your   
					insulin pump, your personal settings, delivery rates, temperature and other factors.   
					With typical pump use [50 U/day using U100 insulin; operating temperature   
					72°F 6°F (22°C 3°C)], pump battery life is about four weeks using alkaline   
					batteries and one week using rechargeable batteries.   
					Chapter 1: Understanding Your New System   
					ACCU-CHEK Spirit Insulin Pump   
					21   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					You must set the battery type to match the kind of battery you are   
					using. If you do not set the correct battery type on your pump, the   
					alert A2: BATTERY LOW may not give you enough time to replace your   
					battery. If you change from using regular batteries to rechargeable   
					batteries, or from rechargeable batteries to regular batteries, you must   
					change the battery type setting in the Standard setup menu (SETUP   
					MENU STANDARD).   
					Always make sure that you remove the battery if your   
					insulin pump is not in use for a long time period in order   
					to preserve the battery lifetime. To prevent water from   
					entering the casing, change the battery only in a dry area.   
					Make sure that the seal of the battery cover is not worn   
					out or missing, and that the battery is properly inserted.   
					! 
					L 
					CAUTION   
					Battery Cover   
					Seal   
					The battery cover seals the battery compartment. Change the battery cover   
					with at least every 4th battery change.   
					• Remove the battery cover only in a dry area to prevent water from   
					getting inside the pump.   
					• Make sure that the seal is not worn or missing and that the battery   
					is properly inserted into the pump.   
					• Tighten or loosen the battery cover using only the ACCU-CHEK Spirit   
					insulin pump battery key. Knives, screwdrivers, or other sharp objects   
					may harm your pump.   
					• Do not over-tighten the battery cover. The battery cover is correctly   
					inserted and tightened when it is even with the pump casing.   
					Chapter 1: Understanding Your New System   
					22   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Battery Key   
					The battery key tightens and loosens the battery cover. It can also loosen the   
					luer-lock connection between the infusion set and adapter if it cannot be   
					loosened by hand. However, never use the battery key to attach or to tighten   
					an infusion set.   
					The battery key has a notch that fits all ACCU-CHEK infusion set luer-lock   
					connectors. The battery key should be replaced if it cracks, becomes worn,   
					or if it is lost.   
					Emergency Kit   
					It is a good idea to carry supplies in case of an emergency so that you can   
					exchange parts if you need to. An emergency kit might contain:   
					• A new ACCU-CHEK infusion set   
					• A new AA alkaline battery   
					• A new ACCU-CHEK 3.15 ml plastic cartridge   
					• A pen or syringe for alternate therapy   
					• A vial of insulin   
					• A filling aid   
					• Treatment for low blood glucose levels   
					• Blood glucose monitoring supplies (such as an ACCU-CHEK blood   
					glucose monitor)   
					• A disinfectant for skin   
					• A new site dressing   
					• The battery key   
					This list is only an example of what an emergency kit might contain. Ask your   
					doctor or healthcare team what supplies you should keep in your personal   
					emergency kit.   
					Chapter 1: Understanding Your New System   
					ACCU-CHEK Spirit Insulin Pump   
					23   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				This page left blank intentionally.   
					Chapter 1: Understanding Your New System   
					24   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Chapter 2: Getting Started   
					Inserting (or Changing) the Battery   
					Your pump turns on when you insert the battery. When you remove the battery,   
					your pump still tracks the time and date for about 1 hour. The pump will always   
					store your pump settings (such as the hourly basal rates, bolus increment and   
					active user menu) and event memory (bolus and alarm history, history of daily   
					insulin totals and Temporary Basal Rates) even if the battery power dies and no   
					matter how long your pump is without power.   
					See the section “Accessories and Disposables – Battery” (page 21) for information   
					on recommended batteries.   
					Keep the Time/Date Settings Correct   
					! 
					L 
					Always check that the time and date of your insulin pump   
					WARNING   
					are set correctly. The time and date are displayed on the   
					STOP screen. Incorrect programming of the time and date   
					may cause incorrect insulin delivery. If you, your doctor,   
					or healthcare team store and analyze your therapy data   
					electronically and the time and date of the devices used   
					are not set identically, the gathered data might not be   
					meaningful. See the section “Setting Time and Date”   
					(page 32) for more information.   
					To prevent water from entering the casing, change the   
					battery only in a dry environment. Make sure that the   
					battery seal is not worn out or missing and that the   
					battery is properly inserted. Tighten or loosen the battery   
					cover with the ACCU-CHEK Spirit insulin pump battery key   
					(knives, screwdrivers or other sharp objects may harm   
					your insulin pump). Do not over-tighten as this may cause   
					damage to the battery cover and to the casing. The battery   
					cover is correctly inserted and tightened when the battery   
					cover is level with the insulin pump casing.   
					! 
					L 
					CAUTION   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					25   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Follow the steps below:   
					1. Press d to move to the STOP YOUR ACCU-CHEK screen.   
					Press f to select and put the pump in STOP   
					mode. Your pump returns to the STOP screen   
					and insulin delivery stops.   
					2. Remove or disconnect your infusion tube set from the headset so that insulin   
					cannot accidentally be delivered.   
					3. Remove the battery cover.   
					Use the battery key to turn the battery cover as   
					indicated by the arrow on the battery cover. Make   
					sure that the opening of the battery compartment   
					and the seal are clean and undamaged. Remove   
					the old battery if necessary.   
					4. Put the battery cover on the positive (+) end of the battery.   
					Hold the cover on the positive (+) end of the battery,   
					and insert the battery negative (-) end first into the   
					battery compartment.   
					Chapter 2: Getting Started   
					26   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				5. Use the cover to gently push the battery all the way in.   
					Turn the battery cover clockwise to tighten it.   
					The battery cover is correctly tightened when   
					the outside of it is even with the casing. Do not   
					over-tighten the battery cover.   
					The pump begins its start-up process.   
					Note:   
					If your pump was in RUN mode before you removed the battery, an   
					error E8: POWER INTERRUPT will occur when a new battery is inserted.   
					Press f twice to confirm and turn off the error. Check the time and   
					date in the STOP screen and correct them, if necessary. See the section   
					“Setting Time and Date” (page 32) for instructions. Put your pump into   
					RUN mode if necessary.   
					Using AA batteries other than those recommended by   
					Disetronic Medical Systems may significantly reduce the   
					! 
					L 
					battery usage time. Batteries other than those recommended   
					may leak and corrode the battery contacts within your   
					insulin pump. Therefore, using batteries not recommended   
					by Disetronic Medical Systems may void the warranty. Do   
					not insert old or used batteries into your insulin pump as   
					this may cause a failure at the insulin pump start-up.   
					CAUTION   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					27   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				System Check Start-Up Process   
					Your pump performs internal tests at start-up when you insert a battery and when   
					you change the cartridge. When you insert a battery the whole start-up process   
					is performed. When only the cartridge is changed, the start-up process begins   
					at the SELF-TEST screen.   
					Do not interrupt the start-up process by pressing keys   
					or by handling the insulin pump. Interruption of the   
					start-up process may lead to malfunctions of your   
					insulin pump.   
					! 
					L 
					WARNING   
					If your insulin pump does not beep and vibrate as   
					described, or if individual numbers, letters, symbols   
					or lines are shown incompletely or not at all, contact   
					Pump Support at 1-800-688-4578.   
					Note:   
					While you are using the pump, its safety system monitors it continuously.   
					If the safety system detects a change from the normal state, an alert   
					(warning instruction) or an error (error message) occurs.   
					1. Put the pump in STOP mode.   
					2. The ACCU-CHEK logo appears.   
					Chapter 2: Getting Started   
					28   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3. SELF-TEST appears.   
					4. Your pump beeps and BEEP TEST appears.   
					Listen for the beeps.   
					If you cannot hear the beeps from your insulin pump, you   
					! 
					may not be notified of changes with your insulin pump in   
					a suitable amount of time. If you cannot hear the beeps,   
					contact Pump Support at 1-800-688-4578.   
					L 
					CAUTION   
					5. Your pump vibrates and VIBRATION TEST appears.   
					Check for vibrations.   
					If you cannot feel the vibration alert, you may not be   
					notified of any changes of your insulin pump in a suitable   
					CAUTION amount of time. If you cannot feel the vibration alert,   
					contact Pump Support at 1-800-688-4578.   
					! 
					L 
					6. The display goes blank. Check that the display is empty. (The backlight turns   
					on during this step.)   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					29   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				7. The display shows a pattern. Check that the pattern is even.   
					8. The display turns black. Check that the display is completely black.   
					9. You will also see an INFORMATION screen.   
					10. If your pump detects an alert or error during the start-up process, it   
					signals the error or alert at this point. See the section “Alerts and Errors”   
					(page 147) for more information.   
					11. A melody signals the end of the start-up process, and the pump shows   
					the STOP screen. The STOP-Warning occurs every minute by giving a   
					long beep and a vibration to remind you that no insulin is being delivered   
					in STOP mode. Press and hold a or s until you hear a melody to turn   
					the STOP-Warning off.   
					Chapter 2: Getting Started   
					30   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					If the voltage of the battery is too low, the start-up procedure   
					cannot finish.   
					Your pump beeps five tones and keeps restarting the start-up   
					procedure until you remove the battery or until the battery is dead.   
					If you discover incomplete characters, numbers, or symbols   
					on the display, put your insulin pump into STOP mode and   
					! 
					L 
					CAUTION   
					perform a system check again by simply removing the battery   
					for a few seconds. When you reinsert the battery the system   
					check will occur. If the incomplete characters, numbers, or   
					symbols remain, your insulin pump may need to be returned   
					since you may not receive the correct information for   
					operating your pump. Put your insulin pump into STOP mode   
					and contact Pump Support at 1-800-688-4578 immediately.   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					31   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Setting Time and Date   
					Be careful to set the time and date correctly, because the basal rates and all   
					history information are stored on the basis of time and date set in the pump.   
					When you travel across time zones, make sure that the time and date are   
					set correctly.   
					The time format can be either American or European/Military Time.   
					American   
					12-hour clock   
					month/day/year   
					European/Military Time   
					24-hour clock   
					day/month/year   
					Incorrect setting of time and date can lead to incorrect   
					insulin delivery. Make sure that the time and date of your   
					insulin pump is programmed correctly to ensure the   
					correct insulin delivery and data memory. Do not make   
					therapy decisions based upon a single result in the   
					insulin pump’s memory.   
					! 
					L 
					WARNING   
					If you, your doctor, or healthcare team reviews your therapy   
					data electronically, it is essential that the time and date   
					of your insulin pump, your blood glucose measurement   
					system (such as an ACCU-CHEK blood glucose meter),   
					the PC, and other devices used are set identically. If they   
					are not, the gathered data might not be meaningful. Check   
					these devices regularly to ensure that the time and date   
					settings are identical.   
					Chapter 2: Getting Started   
					32   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				RUN   
					screen   
					set the time   
					and date   
					set the   
					hour   
					set the   
					minute   
					set the   
					year   
					set the   
					month   
					set the   
					day   
					Follow the steps below to set the Time and Date:   
					1. Press d to move to the SET TIME AND DATE screen.   
					Press f to select.   
					2. The SET HOUR screen displays.   
					Press a or s to set the hour.   
					3. Press d to move to the SET MINUTE screen.   
					Press a or s to set the minutes.   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					33   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				4. Press d to move to the SET YEAR screen.   
					Press a or s to set the year.   
					5. Press d to move to the SET MONTH screen.   
					Press a or s to set the month.   
					6. Press d to move to the SET DAY screen.   
					Press a or s to set the day.   
					7. Press f to save and exit.   
					Chapter 2: Getting Started   
					34   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Your Personal Settings   
					Your new ACCU-CHEK Spirit insulin pump must be   
					! 
					programmed with your personal settings prior to starting   
					insulin pump therapy. Work with your doctor or healthcare   
					team to understand how to use the pump.   
					L 
					WARNING   
					Always check your personal settings before using your   
					new pump in order to avoid incorrect insulin delivery.   
					Record your personal settings from your current insulin   
					pump and ensure that they are correctly programmed   
					into a new ACCU-CHEK Spirit insulin pump. Incorrect   
					programming of your insulin pump may cause incorrect   
					insulin delivery.   
					Do not operate your insulin pump without knowing your   
					personal settings. If you are not sure about your personal   
					settings or you are less experienced, have your doctor or   
					healthcare team check them. Work with your doctor or   
					healthcare team to determine the timing as well as the   
					amount and type of bolus you need to deliver. Make sure   
					you know your personal carbohydrate to insulin ratio and   
					your correction bolus ratio.   
					Change your personal settings only after consulting with your   
					doctor or healthcare team and always follow their instructions.   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					35   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Your Basal Rate Profile   
					Your pump delivers insulin every 3 minutes, in 20 equal doses each hour,   
					24 hours a day. This flow of insulin, measured in international units per hour   
					(U/h), is called the basal rate and is calculated to meet your basic insulin needs.   
					An ACCU-CHEK Spirit insulin pump basal rate profile consists of up to 24   
					different hourly basal rates. Each hourly rate may be changed independently.   
					The total of all 24 hourly basal rates in one basal rate profile is called the   
					(daily) basal rate total.   
					Your basal rate profile should be reviewed:   
					• after any programming changes are made to your pump,   
					• after a new battery is inserted, and   
					• after confirming an error E7: ELECTRONIC ERROR.   
					Chapter 2: Getting Started   
					36   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Programming Your Basal Rate Profile   
					Note:   
					Insulin delivery does not stop or pause during basal rate programming.   
					The insulin amounts and other values used in this user guide are   
					examples only. Your personal values may differ.   
					RUN   
					screen   
					program   
					profile 1   
					daily basal   
					rate total   
					1st hour   
					2nd hour   
					… 
					24th hour   
					Note:   
					For your safety and convenience during programming, your insulin   
					pump returns itself to the RUN or STOP screen if no key is pressed   
					within 20 seconds. The changes you made before your insulin pump   
					“times out” will not be saved.   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					37   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Follow the steps below:   
					1. Press d to move to the PROGRAM BASAL RATE PROFILE 1 screen.   
					2. Press f to select.   
					The daily BASAL RATE TOTAL is displayed.   
					3. Press d to move to the first hour.   
					The first hour, which always begins at midnight,   
					00:00 – 01:00 (12:00 AM – 01:00 AM if the   
					American time format is used), is shown.   
					4. Press a or s or scroll to set the hourly basal rate given by your doctor   
					or healthcare team.   
					5. Press d to move to the next hour.   
					6. Continue using d and the a and s keys to set the hourly basal rates   
					for the remaining hours. This lets you program your individual basal rate   
					profile hour-by-hour. Continue until all 24 hours are programmed.   
					Chapter 2: Getting Started   
					38   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				7. Press f to confirm.   
					The new daily BASAL RATE TOTAL is displayed.   
					Check the new daily BASAL RATE TOTAL.   
					8. Press f to save and exit.   
					If the current active basal rate profile is the same as the new programmed   
					basal rate, it is active immediately.   
					Note:   
					At any screen you have three options to exit:   
					If you want to save the changes   
					• Press f. Check the new daily basal rate total.   
					• Press f. Your pump returns to the RUN screen.   
					If you want to undo the changes   
					• Wait for your ACCU-CHEK Spirit insulin pump to return to the RUN   
					screen (timeout), or   
					• Press d+a simultaneously (exit feature) to exit the function screen.   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					39   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Copying an Hourly Basal Rate   
					If your basal rate is the same for multiple hours in a row, you can save time   
					programming by copying an hourly basal rate to one or more hours.   
					Follow the steps below:   
					1. Press d to move to the PROGRAM BASAL RATE PROFILE 1 screen.   
					2. Press f to select.   
					The daily BASAL RATE TOTAL is displayed.   
					3. Press d to move to the hourly basal rate you want to copy.   
					4. Press a+s simultaneously.   
					A special beep confirms that you are pressing   
					the keys correctly. The same hourly basal rate   
					is copied to the next hour.   
					5. If you want to copy the rate to multiple hours, press a+s repeatedly or   
					press and hold a+s (scroll).   
					Chapter 2: Getting Started   
					40   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				6. Press f to confirm.   
					Check the new daily BASAL RATE TOTAL.   
					7. Press f to save and exit.   
					If the current active basal rate profile is the same as the new programmed   
					basal rate, it is active immediately.   
					Basal Rate and Bolus   
					During normal use, your pump is in RUN mode and insulin is continuously   
					delivered (basal rate).   
					A bolus can be programmed when it is needed. A bolus is the amount of insulin   
					delivered (in addition to the basal rate) to cover the intake of food and correct   
					high blood glucose levels. The bolus amount is determined by your doctor or   
					healthcare team’s guidelines, your blood glucose level, your food intake, and   
					your activity level as well as stress and illness. See the section “Understanding   
					Boluses” (page 75) for more information.   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					41   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Preparing the Cartridge   
					Your insulin pump is designed to use an ACCU-CHEK   
					! 
					3.15 ml plastic cartridge. The cartridge is a sterile product   
					intended for single use only. Sterility is guaranteed for   
					unopened packaging up to the indicated expiration date.   
					L 
					WARNING   
					• Do not use sterile products if the package is damaged.   
					• Do not reuse single use materials. Reusing single use   
					materials might lead to a malfunction of your insulin   
					pump and to incorrect insulin delivery and/or an infection.   
					• Always handle the items with clean hands.   
					• Use specified cartridges only. Using other cartridges   
					can endanger your health and may void the warranty.   
					Chapter 2: Getting Started   
					42   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Filling the Cartridge   
					Note:   
					A filling aid, an optional accessory by Disetronic Medical Systems, helps   
					you fill an empty ACCU-CHEK 3.15 ml plastic cartridge from your vial of   
					insulin. Read the instructions that are included with the filling aid to learn   
					how to use it.   
					To fill a cartridge, have the following materials ready:   
					• A new 3.15 ml plastic cartridge with protective cap, plunger rod,   
					filling needle, and   
					• Insulin at room temperature to minimize formation of air bubbles   
					Follow the steps below:   
					1. Wash your hands.   
					2. Set the insulin vial on a hard, flat surface, such as a table.   
					Clean the rubber membrane of the vial using an   
					antiseptic wipe.   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					43   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3. Remove the cartridge, the cartridge protective cap, and the filling needle   
					with its protective cap from their packaging.   
					4. Place the filling needle with protective cap onto the cartridge tip and make   
					sure the connection is tight.   
					Move the plunger rod back and forth twice in the   
					cartridge to distribute the lubricant.   
					Pull the plunger back to fill the cartridge with air.   
					Plunger   
					Plunger rod   
					5. Remove the protective cap from the needle. Do not touch anything with   
					the needle.   
					Push the tip of the filling needle into the center   
					of the insulin vial’s rubber membrane.   
					Push the plunger rod down so that all air from   
					the cartridge enters the insulin vial.   
					Chapter 2: Getting Started   
					44   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				6. Hold steady pressure against the plunger rod with your thumb, and at the   
					same time turn the insulin vial over so that the filling needle and cartridge   
					are pointing upwards into the insulin vial.   
					Make sure that the tip of the filling needle stays in the insulin.   
					Slowly release the pressure against the plunger rod and   
					let the insulin flow into the cartridge.   
					Do not pull or push the plunger rod while it is moving on   
					its own (this could lead to air bubbles in the insulin).   
					Slowly pull the plunger rod straight down to fill the cartridge.   
					Remove any air bubbles by tapping on the cartridge. Push the   
					air bubbles back into the insulin vial using the plunger rod.   
					The cartridge is full when there are no air bubbles in it and   
					the plunger is at the bottom of the cartridge.   
					7. Remove the filling needle from the insulin vial.   
					8. Place the protective cap on the filling needle.   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					45   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Remove the plunger rod by turning it counter-clockwise   
					out of the plunger.   
					Remove the filling needle with its protective cap from the   
					cartridge by turning counter-clockwise. Dispose of properly.   
					Push the cartridge protective cap tightly onto the cartridge   
					tip until it clicks.   
					The cartridge is now ready to use.   
					Note:   
					Do not pull or push the plunger rod during removal.   
					Air bubbles in the cartridge and infusion set may cause   
					the infusion of air instead of insulin. In this case, your body   
					does not receive the required amount of insulin and an   
					error E4: OCCLUSION may be displayed. Remove bubbles   
					while filling the cartridge and priming the infusion set and   
					without having the infusion set connected to your body.   
					! 
					L 
					CAUTION   
					Cold insulin may release air when warmed. Use only insulin   
					at room temperature when filling the cartridge and priming   
					the infusion set. Inspect the cartridge and the infusion set for   
					air bubbles at least every three hours during the day and   
					before you go to sleep. Remove any air bubbles and change   
					system components if needed.   
					Chapter 2: Getting Started   
					46   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Connecting the Cartridge, Adapter, and Infusion Set   
					Note:   
					Before you change the cartridge, adapter, and infusion set, check   
					that your pump is in good, working order. See the section “System   
					Inspection Checklist” (page 137).   
					Insulin can flow freely from the cartridge or infusion set if the cartridge   
					plunger and the piston rod are not properly connected and the pump   
					is positioned at a higher level than the infusion site. Prevent free flow   
					of insulin by correctly inserting the cartridge: move the piston rod to   
					the proper position and twist the adapter until the cartridge plunger   
					sits flush with the end plate of the piston rod.   
					Have the following materials ready:   
					• A filled ACCU-CHEK 3.15 ml plastic cartridge with a luer-lock connection   
					• An ACCU-CHEK Spirit insulin pump adapter   
					• A new ACCU-CHEK infusion set   
					The cartridge and infusion set are sterile products intended   
					for single use only. Sterility is guaranteed for unopened   
					packaging up to the indicated expiration date. Do not use   
					sterile products if the packaging is damaged.   
					! 
					L 
					WARNING   
					Do not reuse single use materials because this might lead   
					to a malfunction of your insulin pump, to incorrect insulin   
					delivery, and/or an infection. Always handle the items with   
					clean hands. Avoid any contact of the infusion set and the   
					connecting parts of your insulin pump with antiseptics,   
					antibiotic creams, soaps, perfumes, deodorants, body   
					lotions, or any other cosmetics. They may contaminate   
					these parts.   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					47   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Follow the instructions below:   
					Remove the protective cap from the cartridge. Do not touch the cartridge tip.   
					Push the adapter all the way onto the cartridge tip to   
					b 
					the stop (a).   
					a 
					Luer-lock connector   
					Hold the adapter and twist the infusion set luer-lock   
					connection by hand, clockwise into the adapter (b).   
					Carefully prepare a new ACCU-CHEK infusion set for use,   
					priming and insertion, following its instructions for use.   
					Do not over-tighten.   
					To avoid leakage, properly tighten the infusion set to the   
					adapter. Turn the infusion set luer-lock connector until it   
					stops. Do not turn it any further and do not use any auxiliary   
					tools as this may crack the infusion set luer-lock connector   
					and result in leakage.   
					! 
					L 
					CAUTION   
					Note:   
					Your pump cannot detect infusion set leakage. You must check all parts of   
					the infusion set at least every three hours during the day and before you go   
					to sleep. If you detect any loss of insulin and all parts are correctly tightened,   
					immediately replace the leaking component. Because insulin delivery   
					has been interrupted, check your blood glucose level and take appropriate   
					actions according to your doctor’s or healthcare team’s instructions.   
					Chapter 2: Getting Started   
					48   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Inserting the Cartridge   
					Do not perform the CHANGE THE CARTRIDGE function if the   
					! 
					cartridge compartment is not completely dry. During the   
					rewinding of the piston rod, insulin or water might enter your   
					insulin pump and could lead to malfunction. If necessary,   
					use a soft cloth to dry the compartment.   
					L 
					CAUTION   
					3 sec.   
					STOP   
					screen   
					change the   
					cartridge   
					remove the   
					cartridge   
					returning   
					piston rod   
					cartridge volume   
					315 units   
					start-up   
					procedure   
					prime the   
					infusion set   
					Have the following materials ready:   
					• Your insulin pump   
					• A filled ACCU-CHEK 3.15 ml plastic cartridge with a connected   
					ACCU-CHEK Spirit insulin pump adapter   
					• A new ACCU-CHEK infusion set   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					49   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Follow the steps below:   
					1. Press d to move to the STOP YOUR ACCU-CHEK screen.   
					2. Press f to select and put the pump in STOP mode.   
					Your pump displays the STOP screen, and   
					insulin delivery stops.   
					3. Press d to move to the CHANGE THE CARTRIDGE screen.   
					4. Press f to select.   
					Remove the current cartridge, if there is one.   
					5. Press and hold f (for 3 seconds) until you hear a melody.   
					The piston rod begins its return. Do not touch or   
					disturb the piston rod while it is rewinding. Doing   
					so may potentially damage your pump.   
					When the piston rod returns completely, the   
					CARTRIDGE VOLUME screen displays the   
					maximum cartridge volume, 315U.   
					Chapter 2: Getting Started   
					50   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				6. Hold your pump upright. Hold the new filled cartridge so that the connected   
					adapter and infusion tube set point upwards, parallel, and close to the   
					cartridge compartment.   
					Make sure that the bottom edge of the colored   
					part of the adapter is level with the top of the   
					cartridge compartment.   
					Base plate of   
					piston rod   
					Cartridge plunger   
					7. Press a or s to move the plunger rod forward until the end plate of   
					the piston rod is level with the bottom of the cartridge plunger.   
					The cartridge volume decreases accordingly   
					on the display.   
					8. Position your insulin pump upright (with the adapter pointing upwards).   
					Insert the cartridge into the cartridge compartment.   
					9. Twist (do not push) the adapter clockwise until it sits flush with the cartridge   
					compartment.   
					It is not necessary to apply pressure. Twisting will   
					properly position the cartridge. Do not over-tighten   
					the adapter or it will be difficult to remove the   
					cartridge, adapter and cartridge plunger and may   
					potentially cause damage to your pump. The   
					cartridge is correctly inserted when the end plate   
					of the piston rod sits flush against the cartridge   
					plunger. This flush fit is an extra safety measure   
					to prevent the free flow of insulin and to minimize   
					occlusions (blockages).   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					51   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				If the end plate of the piston rod does not sit flush against the cartridge plunger,   
					remove the cartridge from the cartridge compartment. Hold the cartridge   
					and adapter next to the cartridge compartment. Press a or s to move   
					the piston rod forward until its base plate and the cartridge plunger are   
					level. The displayed cartridge volume decreases accordingly. Re-insert the   
					cartridge into the cartridge compartment.   
					– If needed, repeat this step until the cartridge is correctly inserted (the   
					end plate of the piston rod sits flush against the cartridge plunger).   
					10. Press f after correctly inserting the cartridge.   
					Your pump performs a self test. See the section   
					“System Check Start-Up Process” (page 28).   
					11. Check that the infusion set luer-lock connector is still correctly connected   
					to the adapter. Tighten the infusion set by hand clockwise until it sits tightly   
					in the adapter.   
					12. When the system check start-up process is finished, the PRIME THE INFUSION   
					SET screen is displayed.   
					See the section “Priming the Infusion Set” (page 53) for more information.   
					Start from the point in those instructions where the pump screen matches   
					the screen above.   
					Chapter 2: Getting Started   
					52   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Priming the Infusion Set   
					Never prime an infusion set that is connected to your body.   
					You risk uncontrolled insulin delivery into your body. Always   
					follow the instructions for the infusion set you are using.   
					! 
					L 
					WARNING   
					The ACCU-CHEK Spirit insulin pump stops priming after 25 units of insulin. If no   
					insulin has emerged from the tube set needle after priming, repeat the priming   
					function. When insulin comes from the tip of the needle, press any pump key   
					to stop priming. The amount of insulin used in priming is not added to the pump’s   
					history of daily insulin totals.   
					During priming your pump should be upright, with the adapter pointing upwards   
					to help let air bubbles out of the cartridge and the infusion set.   
					Air bubbles in the cartridge and infusion set may cause the   
					infusion of air instead of insulin. If this happens, your body   
					! 
					L 
					will not receive the required amount of insulin. Inspect the   
					cartridge and the infusion set for air bubbles at least every   
					three hours during the day and before you go to sleep.   
					Remove bubbles while filling the cartridge and priming the   
					infusion set and without having the infusion set connected   
					to your body.   
					WARNING   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					53   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				start-up   
					procedure   
					3 sec.   
					stop   
					priming   
					STOP   
					screen   
					prime the   
					infusion set   
					start   
					priming   
					Follow the steps below:   
					1. Press d to move to the STOP YOUR ACCU-CHEK screen.   
					2. Press f to select and put the pump in STOP mode.   
					Your pump displays the STOP screen, and insulin   
					delivery stops.   
					3. Prepare the tubing or cannula/needle for priming per infusion set instructions   
					for use.   
					4. Press d to move to the PRIME THE INFUSION SET screen.   
					Chapter 2: Getting Started   
					54   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				5. Press f to select. The priming amount of 25 units of insulin is shown on   
					the display.   
					6. When START PRIMING is displayed, press and hold f until you hear a   
					melody and the piston rod moves forward.   
					The priming process begins and a count-up is   
					shown on the display. Your pump stops priming   
					after 25 units.   
					When the priming process is finished, your pump   
					returns to STOP.   
					The infusion tubing set is properly primed when   
					there are no air bubbles visible in the tubing and a drop of insulin has come   
					out of the connector or cannula.   
					Stopping the Priming Process   
					You can stop the priming process at any time by   
					pressing any key. Your pump displays the STOP   
					PRIMING screen.   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					55   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Preparing the Infusion Site and Inserting the Infusion Set   
					Always follow the instructions given to you by your doctor or healthcare team   
					and the instructions for the infusion set you are using.   
					Change your infusion set as frequently as recommended by the manufacturer.   
					Infusion Site Selection   
					Your doctor or healthcare team will help you select your infusion site areas   
					according to a planned rotation pattern. Avoid your waistline, bones, scar tissue,   
					recent infusion sites, bruises, and sores. Choose an infusion site at least one inch   
					(2.5 cm) away from the navel and previous infusion sites.   
					Infusion Site Preparation   
					It is important to prepare the infusion site properly to reduce the risks of infection.   
					Your doctor or healthcare team will give you infusion site preparation guidelines.   
					Have the following materials ready:   
					• Your ACCU-CHEK Spirit insulin pump   
					• A disinfectant for your skin   
					• A new ACCU-CHEK infusion set   
					Follow the steps below:   
					1. Make sure your pump is correctly set up (including cartridge, adapter and   
					infusion set) and is programmed with your personal settings.   
					Chapter 2: Getting Started   
					56   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				2. Wash your hands thoroughly.   
					3. Disinfect the infusion site and let it dry completely.   
					4. Make sure your pump is in STOP mode.   
					5. Insert your ACCU-CHEK infusion set according to the instructions for its use.   
					Chapter 2: Getting Started   
					ACCU-CHEK Spirit Insulin Pump   
					57   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					If you use a soft cannula-type infusion set, you must proceed with a bolus   
					to fill the air space in the cannula after removing the introducer needle   
					according to the instructions for the infusion set you are using. Failure to   
					do so could result in missed insulin dosing. See the section “Programming   
					a Standard Bolus” (page 77) for more information.   
					Inspect your infusion site at least once or twice each day for   
					irritation or infection. Signs of infection may include, but   
					are not limited to, pain, lumps, redness, heat, or discharge.   
					! 
					L 
					CAUTION   
					If you see redness or swelling, change your infusion headset   
					and infusion site immediately and contact your doctor or   
					healthcare team.   
					Change your infusion headset according to the instructions   
					for the infusion set you are using and your doctor or   
					healthcare team’s recommendations.   
					If you have read and followed all instructions, cautions, warnings, and notes   
					found in this chapter, your pump should now be ready to use. However, please   
					read the complete manual before starting.   
					Chapter 2: Getting Started   
					58   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Chapter 3: Using Your Pump   
					Before Using   
					Training and using your insulin pump requires the support of   
					an experienced doctor or healthcare team. Regular visits   
					to your doctor or healthcare team are absolutely essential   
					during insulin pump therapy.   
					! 
					L 
					CAUTION   
					Always follow the instructions given by your doctor or   
					healthcare team.   
					Successful insulin pump therapy requires you to check your blood glucose   
					levels often. It is a good idea to check your blood glucose (using an ACCU-CHEK   
					blood glucose monitor) at least four times a day, or as often as you are   
					directed by your doctor or healthcare team.   
					Wearing Your Pump   
					When wearing your insulin pump make sure to prevent   
					any contact with objects that may damage or accidentally   
					press the keys of your insulin pump (e.g., keychains, keys,   
					pocket knives, and coins).   
					! 
					L 
					CAUTION   
					Attach your pump safely to your body or clothes to prevent damage. Special   
					carrying systems let you carry your pump on or under your clothes. All specified   
					accessories have been tested and approved for use with your ACCU-CHEK Spirit   
					insulin pump. The proper functioning of your pump can only be guaranteed if you   
					use it with ACCU-CHEK Spirit insulin pump accessories.   
					See Disetronic Medical Systems sterile products and accessories catalogs and   
					brochures for more information. Or, contact Pump Support at 1-800-688-4578.   
					Chapter 3: Using Your Pump   
					ACCU-CHEK Spirit Insulin Pump   
					59   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Wear your pump under your clothes or directly on your body in cold or rainy   
					weather. If you have concerns about the weather affecting your pump,   
					check the operating conditions of your pump and contact Pump Support at   
					1-800-688-4578. See the section “Appendix A: Technical Data” (page 165) for   
					more information on operating conditions.   
					Do not place your insulin pump in direct sunlight. Overheating   
					of the insulin and your insulin pump must be avoided. Protect   
					your insulin pump from direct exposure to cold wind and   
					temperatures over 104°F (40°C) and below 41°F (5°C). These   
					conditions may damage the insulin and the electronics of   
					your insulin pump and may cause a malfunction of the battery.   
					! 
					L 
					CAUTION   
					See the instructions for the insulin you are using to determine   
					the acceptable temperature range.   
					Chapter 3: Using Your Pump   
					60   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				When Not to Wear Your Pump   
					Electromagnetic Fields and Hazardous Areas   
					Do not use your pump around electromagnetic fields such as radar or antenna   
					installations, high-voltage sources, X-Ray sources, MRI, CAT scan, and all other   
					sources of electrical current as they may cause your pump not to work. Insulin   
					delivery may stop and an error E7: ELECTRONIC ERROR may occur.   
					Always stop and remove your pump before you enter these areas.   
					See “Appendix A: Technical Data” (page 165) for more information about   
					electromagnetic fields.   
					Your insulin pump has not been tested with cardiac   
					pacemakers. Do not use the ACCU-CHEK Spirit insulin   
					pump if you have a cardiac pacemaker.   
					! 
					L 
					WARNING   
					Your pump has been tested and complies with the regulations regarding   
					unintentional electromagnetic interference. Security systems in airports as   
					well as anti-theft monitoring devices, such as those in department stores,   
					should not affect the functioning of your pump.   
					Because there are so many devices, such as cellular phones, that give out   
					electromagnetic radiation, the influence of some such devices on your pump   
					cannot be ruled out. It is a good idea to keep your pump at least 4 inches   
					(10 cm) away from such devices while the other devices are turned on.   
					Do not use your pump in hyperbaric chambers and in hazardous areas of any   
					kind, such as areas where explosive or flammable gases or vapors could exist.   
					This could affect insulin delivery and/or lead to harm.   
					Your pump is designed to work in normal barometric conditions (from 700   
					to 1,060 mbar).   
					Chapter 3: Using Your Pump   
					ACCU-CHEK Spirit Insulin Pump   
					61   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Do not exceed 10,000 feet (3,000 meters) above sea level. Your pump has   
					not been tested for use above this level.   
					Always stop and remove your pump before you enter these areas. If you   
					have questions, contact Pump Support at 1-800-688-4578.   
					Sports and Exercise   
					Exercise is an important part of diabetes management. Disetronic Medical   
					Systems designed the ACCU-CHEK Spirit insulin pump to let you safely perform   
					many sporting activities while you wear your pump. It is important that you   
					protect your pump before you start exercise or sports activities. Do not wear   
					your pump during sports that involve body contact such as boxing, football,   
					or hockey since rough contact could damage your pump.   
					Consult our sterile products brochures and accessories catalog for more   
					information regarding carrying systems or contact Pump Support at   
					1-800-688-4578 for more information.   
					Traveling   
					Ask your doctor or healthcare team about any special actions you need to   
					take before you travel. Take extra blood-glucose testing and pump supplies   
					when you travel and find out where to get supplies while you are traveling.   
					When you travel across time zones, make sure that the time and date on   
					your pump are set correctly to the local time.   
					The ACCU-CHEK insulin pump has not been tested at altitudes in excess of   
					10,000 feet (3,000 meters) above sea level. If your travels take you to an area   
					in excess of 10,000 feet (3,000 meters) above sea level, stop and remove your   
					pump. This does not apply to use in airplanes with pressurized cabins.   
					If you have questions, contact Pump Support at 1-800-688-4578.   
					Chapter 3: Using Your Pump   
					62   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Your Insulin Pump and Water   
					Avoid deliberate contact with water. Check daily that your   
					! 
					insulin pump is not chipped, cracked, or damaged in any   
					way and that the battery cover and the adapter are correctly   
					closed. In the presence of chips and cracks, water, dust,   
					insulin, or other foreign substances may enter your insulin   
					pump and lead to malfunction. Prior to contact with water,   
					disconnect and take off your insulin pump.   
					L 
					CAUTION   
					Disconnect and remove your pump before entering a bathtub, Jacuzzi or   
					whirlpool, shower, or swimming pool. Avoid exposing your pump to high   
					humidity, such as in a sauna, as this could damage your pump. To make it   
					easy to remove your pump, use disconnectable ACCU-CHEK infusion sets.   
					Consult your doctor or healthcare team to find out how long your insulin pump   
					therapy can safely be interrupted.   
					Accidental Contact with Water   
					Your insulin pump is protected against accidental short-term water contact.   
					Examples of accidental short-term water contact include:   
					• rain, snow;   
					• water that splashes while you are biking, jogging, or hiking;   
					• accidental brief immersion into the sink or bathtub.   
					Chapter 3: Using Your Pump   
					ACCU-CHEK Spirit Insulin Pump   
					63   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				What to Do After Any Water Contact   
					Put your pump into STOP mode. Disconnect the pump from your body before   
					inspecting it. Use a soft cloth to dry the outside of the pump. Check the battery   
					compartment and the cartridge compartment for water. If the battery compartment   
					or cartridge compartment are wet, turn your pump upside down to let the water   
					run out and let it air dry. Do not use warm air such as a hair dryer to dry the pump   
					as this could damage the pump’s electronics. Do not re-insert the battery or the   
					cartridge until the compartments are completely dry.   
					See the section “Cleaning Your ACCU-CHEK Spirit Insulin Pump” (page 141)   
					for more information.   
					Other Liquids   
					You do not have to worry about contact between your pump and:   
					• sweat   
					• saliva   
					You must check your pump immediately after contact with other liquids or   
					chemicals such as:   
					• cleaning solutions   
					• alcohol   
					• beverages   
					See the section “Cleaning Your ACCU-CHEK Spirit Insulin Pump” (page 141)   
					for more information.   
					Avoid any contact of your insulin pump, the infusion set,   
					and the connecting parts of your insulin pump with health   
					and beauty items (for example, antiseptics, antibiotic   
					creams, soaps, perfumes, deodorants, body lotions, or any   
					other cosmetics). These items may discolor your pump or   
					fog your display.   
					! 
					L 
					CAUTION   
					Chapter 3: Using Your Pump   
					64   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Starting Insulin Delivery   
					STOP   
					screen   
					start your   
					ACCU-CHEK   
					RUN   
					screen   
					Follow the steps below:   
					1. Start from the STOP screen.   
					2. Press d to move to the START YOUR ACCU-CHEK screen.   
					3. Press f to select. Your pump shows the RUN screen.   
					Insulin delivery begins within 3 minutes.   
					Note:   
					Check the amount of insulin remaining in the cartridge at least once a   
					day. If you press the f key from the RUN or STOP screen, your pump   
					displays the remaining cartridge content in units. Before going to sleep,   
					make sure that:   
					• the cartridge contains enough insulin to last through the night, and   
					• the time and date are set correctly.   
					Chapter 3: Using Your Pump   
					ACCU-CHEK Spirit Insulin Pump   
					65   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Stopping Insulin Delivery   
					When your insulin pump is in STOP mode, it does not deliver   
					any insulin. You will need to put your insulin pump into RUN   
					mode to continue insulin delivery.   
					! 
					L 
					CAUTION   
					There are functions and actions, for your safety, that you can only perform   
					when the pump is in STOP mode. These functions and actions include:   
					• changing the cartridge   
					• connecting and disconnecting the adapter and/or infusion set, and   
					• priming the infusion set   
					Follow the steps below:   
					1. Press d to move to the STOP YOUR ACCU-CHEK screen.   
					2. Press f to select.   
					Your pump returns to the STOP screen and   
					insulin delivery stops.   
					3. Press f (QUICK INFO screen) to review activated functions and features   
					(like the beep and vibration alarm signals) indicated by icons and messages   
					on the screen.   
					Chapter 3: Using Your Pump   
					66   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					When an Extended Bolus, a MultiWave Bolus, and/or a Temporary Basal   
					Rate (TBR) is/are active and you put your pump into STOP mode, an alert   
					A6: TBR CANCELLED and/or an alert A8: BOLUS CANCELLED occur(s).   
					Press f twice for each appearing alert to confirm and turn off. See the   
					sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8: BOLUS   
					CANCELLED” (page 153) for more information.   
					STOP-Warning   
					The STOP-Warning is turned on by the pump each time your pump switches from   
					RUN mode to STOP mode, and when a battery is inserted. The STOP-Warning   
					is a long beep and a vibration occurring every minute. It warns you that insulin   
					delivery has stopped. If you are aware of the situation and don’t wish to be   
					reminded, you can turn off the STOP-Warning.   
					To Turn Off the STOP-Warning   
					Press either a or s for 3 seconds to turn off the STOP-Warning. A melody   
					plays indicating the STOP-Warning is turned off.   
					Note:   
					The volume of the STOP-Warning cannot be changed. It is independent of   
					the programmed beep volume.   
					Chapter 3: Using Your Pump   
					ACCU-CHEK Spirit Insulin Pump   
					67   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Interruption of Insulin Pump Therapy   
					Consult with your doctor or healthcare team about when and for how long your   
					insulin pump therapy can be interrupted.   
					Measure your blood glucose level regularly during any insulin interruption. Use   
					a syringe or insulin pen to inject insulin according to the instructions of your   
					doctor or healthcare team.   
					Since only rapid-acting insulin is used for insulin pump   
					therapy, there is only a small insulin reserve in the   
					body. If insulin is interrupted for any reason, such as:   
					! 
					L 
					WARNING   
					• if you stop the pump   
					• there is a technical problem with the pump   
					• there is leakage in the cartridge   
					• there is occlusion in the infusion tube set or in the   
					infusion headset   
					• the cannula has slipped out of the infusion site   
					you must be ready to replace the missing insulin   
					immediately.   
					Always carry spare sterile products and accessories   
					(infusion set, insulin cartridge, batteries), an insulin   
					pen/syringe, and insulin.   
					Without insulin, diabetic ketoacidosis may develop   
					and may require in-patient hospital treatment.   
					Chapter 3: Using Your Pump   
					68   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Short Interruption   
					1. Put your pump in STOP mode.   
					2. Remove or disconnect the infusion tube set from the infusion headset. When   
					using a disconnectable infusion set, disconnect it at your infusion site and   
					use the protective covers that came with the infusion set.   
					Resuming Insulin Pump Therapy   
					1. Attach a new ACCU-CHEK infusion set and prime it before inserting it.   
					2. Follow the instructions to prepare the infusion set you are using, and put your   
					pump into RUN mode.   
					Measure your blood glucose level within 2 hours of restarting pump therapy to   
					check the function of the system and to ensure insulin delivery.   
					Interruption of Use of Your Insulin Pump for a Long Period of Time   
					There may be times when you may need to interrupt your pump therapy for more   
					than one day. Follow the steps below.   
					Contact your doctor or healthcare team for an alternate   
					therapy plan when you interrupt the insulin pump therapy   
					for a long period of time.   
					! 
					L 
					CAUTION   
					1. Put your pump into STOP mode.   
					2. Remove the cartridge, adapter, and infusion set.   
					3. Reinsert the adapter.   
					4. Remove the battery and reinsert the battery cover.   
					5. Store your pump properly. See the section “Storing Your ACCU-CHEK Spirit   
					Insulin Pump” (page 140) for more information.   
					Chapter 3: Using Your Pump   
					ACCU-CHEK Spirit Insulin Pump   
					69   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Changing the Infusion Set   
					When you change the infusion set, keep the following points and warnings in mind:   
					• Infusion sets are sterile products intended for single use only. Sterility is   
					guaranteed for unopened packaging up to the indicated expiration date.   
					Do not use sterile products if the package is damaged.   
					• Properly tighten the infusion set to the adapter to avoid leakage. Turn the   
					infusion set luer-lock connector until it stops. Do not turn it any further and   
					do not use any auxiliary tools as this may crack the infusion set luer-lock   
					connector and result in leakage.   
					• Your pump cannot detect infusion set leakage. You must inspect all parts   
					of your infusion set at least every three hours during the day and before   
					you go to sleep. If there seems to be any loss of insulin and all parts are   
					correctly tightened, immediately replace the leaking component.   
					Immediately check your blood glucose level, because insulin delivery has   
					been interrupted. Take appropriate actions according to your doctor’s or   
					healthcare team’s instructions.   
					• Always disconnect your infusion set before changing a cartridge or   
					priming. You risk uncontrolled insulin delivery into your body. Always   
					follow the instructions for the infusion set you are using.   
					Have the following materials ready:   
					• Your ACCU-CHEK Spirit insulin pump   
					• A disinfectant for your skin   
					• A new ACCU-CHEK infusion set   
					Chapter 3: Using Your Pump   
					70   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Follow the steps below:   
					1. Wash your hands thoroughly.   
					2. Make sure your pump is in STOP mode.   
					3. Remove the infusion tube set from the headset.   
					4. Remove the infusion tube set from the adapter and dispose of it properly.   
					Note:   
					The battery key has a notch that fits   
					all ACCU-CHEK infusion set luer-lock   
					connectors. You may use the battery   
					key to loosen the luer-lock connector   
					between your infusion set and adapter   
					if it cannot be loosened by hand.   
					5. Carefully prepare a new ACCU-CHEK infusion set for use.   
					6. Turn the ACCU-CHEK infusion set by hand clockwise into the adapter.   
					7. Properly tighten the luer-lock connector by hand.   
					Chapter 3: Using Your Pump   
					ACCU-CHEK Spirit Insulin Pump   
					71   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				8. Prime the infusion set following all the instructions and warnings found in   
					the section “Priming the Infusion Set” (page 53).   
					9. Insert a new infusion set following the instructions found in the section   
					“Preparing the Infusion Site and Inserting the Infusion Set” (page 56).   
					10. Remove the old infusion headset and dispose of it properly.   
					11. Place the pump in RUN mode. See the section “Starting Insulin Delivery”   
					(page 65).   
					Note:   
					If you use a soft cannula-type infusion set, you must proceed with a   
					bolus to fill the air space in the cannula after removing the introducer   
					needle according to the instructions for the infusion set you are using.   
					Failure to do so could result in missed insulin dosing. See the section   
					“Programming a Standard Bolus” (page 77) for more information.   
					Inspect your infusion site at least once or twice each day for irritation   
					or infection. Signs of infection may include, but are not limited to, pain,   
					lumps, redness, heat, or discharge.   
					If you see redness or swelling, change the infusion headset and the   
					infusion site immediately and contact your doctor or healthcare team.   
					Change the infusion headset according to the instructions for the infusion   
					set you are using and your doctor or healthcare team’s recommendations.   
					Chapter 3: Using Your Pump   
					72   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Changing the Cartridge and Infusion Set   
					Have the following materials ready:   
					• Your ACCU-CHEK Spirit insulin pump   
					• A disinfectant for your skin   
					• A new ACCU-CHEK infusion set   
					• A filled ACCU-CHEK 3.15 ml plastic cartridge with a   
					luer-lock connection   
					• A new adapter (must be changed at least with every 10th cartridge)   
					1. Wash your hands thoroughly.   
					2. Make sure your insulin pump is in STOP mode.   
					3. Remove your infusion tube set from the headset.   
					4. Remove the infusion tube set from the adapter and dispose of it properly.   
					Note:   
					The battery key has a notch that fits   
					all ACCU-CHEK infusion set luer-lock   
					connectors. You may use the battery   
					key to loosen the luer-lock connector   
					between your infusion set and adapter   
					if it cannot be loosened by hand.   
					Chapter 3: Using Your Pump   
					ACCU-CHEK Spirit Insulin Pump   
					73   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				5. Hold your ACCU-CHEK Spirit insulin pump   
					with the adapter pointing downwards.   
					6. Remove the adapter and cartridge from your pump by unscrewing the   
					adapter from the pump casing. Pulling the adapter from the pump casing   
					without completely unscrewing it may result in the plunger staying attached   
					to the piston rod. If this should happen, please call Pump Support at   
					1-800-688-4578.   
					7. Remove the cartridge from the adapter and dispose of the cartridge properly.   
					8. Hold the adapter up to the light to check for signs of wear or dirt (especially   
					on the seals inside and outside of the adapter). It is recommended that you   
					replace the adapter at least every 10th insulin cartridge change.   
					9. If needed, clean the adapter using water and dry it. If it shows signs of   
					wear or dirt, replace it immediately.   
					10. Install a new cartridge and a new infusion set. See the following sections   
					for instructions: “Changing the Infusion Set” (page 70), “Preparing the   
					Cartridge” (page 42), “Connecting the Cartridge, Adapter, and Infusion Set”   
					(page 47), “Inserting the Cartridge” (page 49), “Priming the Infusion Set”   
					(page 53), “Preparing the Infusion Site and Inserting the Infusion Set”   
					(page 56), and “Starting Insulin Delivery” (page 65).   
					Note:   
					Check the amount of insulin remaining in the cartridge at least once   
					a day. Press the f key from the RUN or STOP screen to display   
					the remaining cartridge content in units. Before going to sleep, make   
					sure that:   
					• the cartridge contains enough insulin to last through the night, and   
					• that the time and date are set correctly.   
					Chapter 3: Using Your Pump   
					74   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Chapter 4: Boluses and Temporary Basal Rates   
					Understanding Boluses   
					A bolus is the amount of insulin delivered (in addition to the basal rate) to cover   
					the intake of food and correct high blood glucose levels. The bolus type and   
					amount is determined by your doctor or healthcare team’s guidelines, your   
					blood glucose level, your food intake, your health condition, and your activity   
					level. Discuss the timing and the amount and type of bolus you need with your   
					doctor or healthcare team. You can review the time, date, and amount of the   
					last 30 boluses in the bolus history. See the section “Review the Bolus History”   
					(page 107) for more information.   
					Your pump can deliver three types of boluses:   
					Standard Bolus   
					Extended Bolus   
					MultiWave Bolus   
					fast delivery   
					delivery over a programmed period of time   
					combines fast delivery with delivery over a   
					programmed period of time   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					75   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				• Incorrect programming of your insulin pump may cause   
					inappropriate insulin delivery.   
					• Your insulin pump must be programmed with your   
					personal settings prior to starting insulin pump therapy.   
					• Do not operate your insulin pump without knowing your   
					personal settings.   
					! 
					L 
					WARNING   
					• If you are not sure about your personal settings or you   
					are less experienced, have your doctor or healthcare   
					team check your personal settings.   
					• Work with your doctor or healthcare team to determine   
					the timing, amount, and type of bolus you need to deliver.   
					• Make sure you know your personal carbohydrate to   
					insulin ratio and your correction bolus ratio.   
					Chapter 4: Boluses and TBRs   
					76   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Programming a Standard Bolus   
					There are two ways to program a Standard Bolus:   
					“Quick” Standard Bolus   
					“Scroll” Standard Bolus   
					Using a and s on your   
					insulin pump, with bolus increments   
					determined by the user   
					Menu-guided using d and f   
					with a and s for   
					programming of the amount   
					The bolus amount per delivery is limited to 25 units using U100 insulin.   
					After you program a Standard Bolus there is a delay of 5 seconds before the   
					bolus delivery starts. This delay lets you cancel the bolus before delivery starts   
					by pressing the a or s key. Alert A8: BOLUS CANCELLED will occur. Press   
					f twice to confirm and turn off the alert. See the section “Alert A8: BOLUS   
					CANCELLED” (page 153) for more information.   
					Programming a “Quick” Standard Bolus   
					The following process involves only the a and s keys. You can start with   
					either key. The recommended steps below are those described with the first   
					key appearing without parenthesis, such as a. For the alternative step press   
					the key in the parenthesis, such as (s).   
					Note:   
					Because your pump beeps and vibrates to guide you safely through the   
					programming, you may program the “Quick” STANDARD Bolus using   
					one finger. After the bolus is programmed, you can count the beeps and   
					vibrations of your pump to verify the correct programming.   
					The bolus increment for the “Quick” Standard Bolus is initially set to   
					0.5 units.   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					77   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				3 sec.   
					RUN   
					screen   
					bolus   
					amount   
					5 sec. delay   
					bolus   
					delivery   
					if O unit   
					3 sec.   
					confirmation   
					of amount   
					5 sec. delay   
					A8: bolus cancelled   
					Follow the steps below:   
					1. From the RUN screen press and hold s (a) until you hear a beep   
					sequence and feel a vibration.   
					This activates the “Quick” STANDARD   
					BOLUS function.   
					Hint:   
					Start programming with the a key in areas of poor lighting, as this   
					will also turn on the backlight.   
					2. Press a (s) repeatedly until the bolus amount is reached.   
					Each press of the a (s) key adds one bolus   
					increment to the bolus amount. Your pump   
					beeps and vibrates simultaneously once for   
					each programmed bolus increment.   
					Five seconds after the last press of a (s), your pump confirms the total   
					bolus amount with one beep and vibration for each bolus increment programmed.   
					The Standard Bolus symbol ( ) blinks for 5 seconds (the bolus delivery   
					start delay).   
					After this bolus delivery start delay, the pump beeps three times and then   
					begins to deliver the total bolus amount programmed. The countdown of the   
					remaining bolus amount appears on the display.   
					Chapter 4: Boluses and TBRs   
					78   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				The bolus increment that can be programmed into your   
					insulin pump determines the Standard Bolus amount   
					using the a and s keys of your insulin pump.   
					Incorrect programming of the bolus increment may   
					cause inappropriate insulin delivery. See the section   
					“Bolus Increment” (page 131) for more information.   
					! 
					L 
					WARNING   
					Cancelling a “Quick” Standard Bolus   
					– during programming (the bolus amount blinks):   
					• Press the key s or a you used to enter the “Quick” Standard   
					Bolus programming option. This will reset the bolus amount to 0.0   
					units. A melody will play. If no new bolus is programmed, your pump   
					returns to the RUN screen and beeps three times after five seconds.   
					No bolus is delivered.   
					– during confirmation (when the beeps and vibrations occur), or during the   
					start delay ( blinks):   
					• Press s or a. A melody plays. Your pump returns to the RUN   
					screen. An alert A8: BOLUS CANCELLED occurs. Press f twice to   
					confirm and turn off the alert. See the section “Alert A8: BOLUS   
					CANCELLED” (page 153) for more information.   
					No bolus is delivered.   
					– during bolus delivery (countdown of bolus amount):   
					• Press and hold s or a for 3 seconds until you hear a melody.   
					An alert A8: BOLUS CANCELLED occurs. Press f twice to confirm   
					and turn off the alert.   
					Bolus delivery is interrupted.   
					See the section “Alert A8: BOLUS CANCELLED” (page 153) for   
					more information.   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					79   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				You can view the actual bolus amount delivered prior to the cancellation by   
					looking at the bolus history. See the section “Review the Bolus History” (page   
					107) for more information.   
					Make sure that the cancellation was intended, and then program a new Standard   
					Bolus, if needed.   
					Programming a Menu-Guided “Scroll” Standard Bolus   
					Regardless of the bolus increment setting, the bolus increments for the “Scroll”   
					Standard Bolus are fixed to 0.1 unit. You can program this bolus in the STANDARD   
					BOLUS screen by holding the a or s key (scrolling) until the bolus amount   
					is displayed.   
					Follow the steps below:   
					1. Press d to move to the STANDARD BOLUS screen.   
					2. Press f to select. The BOLUS AMOUNT screen appears.   
					3. Press or scroll a to increase or s to decrease the bolus amount.   
					4. Press f to confirm the bolus amount.   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					80   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				The Standard Bolus symbol ( ) blinks for five seconds (the bolus delivery   
					start delay).   
					After this bolus delivery start delay, your pump beeps three times and begins to   
					deliver the total bolus amount programmed. The countdown of the remaining   
					bolus amount appears on the display.   
					Cancelling a Menu-Guided “Scroll” Standard Bolus   
					– during programming (the bolus amount blinks):   
					You have three options for cancelling a “Scroll” Standard Bolus during   
					programming:   
					• If you do not press any key for 20 seconds, your pump returns to the   
					RUN screen (timeout).   
					• Exit to the STANDARD BOLUS menu by pressing d+a simultaneously.   
					• Set the bolus amount to 0.0 units. Press f to exit.   
					No bolus is delivered.   
					– during the start delay ( is blinking):   
					• Press s or a. A melody plays. Your pump returns to the RUN screen.   
					An alert A8: BOLUS CANCELLED occurs. Press f twice to confirm   
					and turn off the alert.   
					No bolus is delivered.   
					Please refer to section “Alert A8: BOLUS CANCELLED” (page 153) for more   
					information.   
					– during bolus delivery (countdown of bolus amount):   
					• Press and hold s or a for 3 seconds until you hear a melody.   
					An alert A8: BOLUS CANCELLED occurs. Press f twice to confirm   
					and turn off the alert.   
					Bolus delivery is interrupted.   
					Please refer to section “Alert A8: BOLUS CANCELLED” (page 153) for   
					more information.   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					81   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				You can view the actual bolus amount delivered prior to the cancellation   
					by looking at the bolus history. Please refer to the section “Review the   
					Bolus History” (page 107) for more information.   
					Make sure that the cancellation was intended and then program a new Standard   
					Bolus, if needed.   
					Chapter 4: Boluses and TBRs   
					82   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Programming an Extended Bolus   
					The Extended Bolus is available in the ADVANCED and CUSTOM user menus. The   
					Extended Bolus function lets you program a bolus to be delivered over a period of   
					time. This function is helpful during long meals, dinners, or receptions, or when   
					you have meals that are digested slowly. The Extended Bolus may also be   
					appropriate for people who have gastroparesis (delayed digestion).   
					The duration of bolus delivery can be programmed in 15-minute intervals up to   
					24 hours and begins immediately after confirmation.   
					Note:   
					You can add a Standard Bolus to an ongoing Extended Bolus. If you   
					cancel the Standard Bolus, the Extended Bolus continues to run.   
					During the delivery of an Extended Bolus, the programming of another   
					Extended Bolus or a MultiWave Bolus is blocked. If you want to program   
					a different Extended Bolus, put the pump in STOP mode to cancel the   
					current bolus and to program a new one.   
					RUN   
					screen   
					Extended   
					bolus   
					bolus   
					amount   
					bolus   
					duration   
					stop your   
					ACCU-CHEK   
					A8: bolus cancelled   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					83   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Follow the steps below:   
					1. Press d to move to the EXTENDED BOLUS screen.   
					2. Press f to select. The BOLUS AMOUNT screen appears.   
					3. Press a to increase or s decrease the bolus amount.   
					The duration of the last Extended Bolus   
					delivered appears, or when using your pump   
					for the first time, a default value appears.   
					4. Press d to move to the BOLUS DURATION screen.   
					Press a to increase or s decrease the   
					bolus duration.   
					Note:   
					Press d to loop between programming the bolus amount and   
					bolus duration.   
					5. Press f to confirm the bolus amount and duration. You will hear a melody   
					when the bolus delivery begins within the next 3 minutes.   
					Chapter 4: Boluses and TBRs   
					84   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					At any function screen you have three options to exit:   
					If you want to deliver the bolus   
					• Press f. Your pump returns to the RUN screen.   
					If you want to cancel the bolus   
					• Wait for your pump to return to the RUN screen (timeout), or   
					• Press d+a simultaneously (exit feature) to exit the function screen.   
					• Set the bolus amount to 0.0 units and press f.   
					Note:   
					During the entire bolus delivery, the remaining time and amount of the   
					Extended Bolus and the current hourly basal rate are displayed in the   
					RUN screen.   
					Note:   
					If a Temporary Basal Rate is active at the same time as the Extended Bolus,   
					the remaining time and amount of the bolus and the amount of the   
					increased or decreased hourly basal rate are displayed in the RUN screen.   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					85   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Cancelling an Extended Bolus   
					– during programming (the bolus amount or bolus duration blinks):   
					You have three options for cancelling an Extended Bolus during   
					programming:   
					• If you do not press any key for 20 seconds, your pump returns to the   
					RUN screen (timeout).   
					• Exit to the EXTENDED BOLUS menu by pressing d+a simultaneously.   
					• Set the bolus amount to 0.0 units and press f.   
					No bolus is delivered.   
					– during bolus delivery:   
					• If the Extended Bolus delivery has begun, it can be cancelled by putting   
					your pump into STOP mode. This cancels the bolus delivery and an alert   
					A8: BOLUS CANCELLED occurs. Press f twice to confirm and turn off   
					the alert. See the section “Alert A8: BOLUS CANCELLED” (page 153) for   
					more information.   
					Bolus delivery is interrupted.   
					The bolus amount delivered up to the cancellation can be reviewed in   
					the BOLUS HISTORY screen. See the section “Review the Bolus   
					History” (page 107) for more information.   
					When your insulin pump is in STOP mode, it does not deliver   
					any insulin. Put the pump in RUN mode to continue the   
					CAUTION insulin delivery.   
					! 
					L 
					Chapter 4: Boluses and TBRs   
					86   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					If a Temporary Basal Rate is active at the same time, it is also cancelled   
					when your pump is put into STOP mode. An alert A8: BOLUS CANCELLED   
					and an alert A6: TBR CANCELLED occur.   
					Press f twice to confirm and turn off the first alert. The second alert   
					appears on the display. Press f twice to confirm and turn off the second   
					alert. Both alerts are recorded in the alarm history.   
					See the sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8:   
					BOLUS CANCELLED” (page 153) for more information.   
					Make sure that the cancellation was intended and then program a new   
					Extended Bolus (and/or Temporary Basal Rate), if needed.   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					87   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Programming a MultiWave Bolus   
					The MultiWave Bolus function is available in the ADVANCED user menu. It is   
					designed to better simulate a body’s insulin delivery. It combines an immediate   
					bolus delivery followed by an Extended Bolus delivery. This function can be   
					helpful when you have meals that include both rapid and slowly absorbed   
					carbohydrates.   
					The duration of bolus delivery can be programmed in 15-minute intervals for   
					up to 24 hours.   
					Note:   
					• The MultiWave Bolus is available in the ADVANCED user menu.   
					• You can add a Standard Bolus to an ongoing MultiWave Bolus. If you   
					cancel the Standard Bolus, the MultiWave Bolus remains running.   
					• During the delivery of a MultiWave Bolus, the programming of a   
					second MultiWave Bolus or an Extended Bolus is blocked. If you   
					need a different MultiWave Bolus, put your pump in STOP mode   
					to cancel the current bolus and program a new one.   
					Chapter 4: Boluses and TBRs   
					88   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				RUN   
					screen   
					MultiWave   
					bolus   
					bolus   
					amount   
					bolus   
					delivery   
					3 sec.   
					immediate   
					bolus   
					or   
					stop your   
					ACCU-CHEK   
					bolus   
					duration   
					5 sec. delay   
					A8: bolus cancelled   
					Follow the steps below:   
					1. Select the ADVANCED user menu or turn on the MULTIWAVE BOLUS menu   
					choice in the CUSTOM user menu. Also, your pump must be in RUN mode.   
					2. Press d to move to the MULTIWAVE BOLUS screen.   
					3. Press f to select.   
					The BOLUS AMOUNT screen appears.   
					4. Press a or s to increase or decrease the total bolus amount.   
					The first time you press a, the bolus duration of   
					the last MultiWave Bolus appears, except when   
					you are using your pump for the first time, in which   
					case, a default value appears.   
					As you increase or decrease the total bolus amount, the suggested immediate   
					bolus amount increases or decreases proportionally.   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					89   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				5. Press d to move to the IMMEDIATE BOLUS screen.   
					Press a or s to correct and set the immediate   
					bolus amount.   
					6. Press d to move to the BOLUS DURATION screen.   
					Press a or s to correct and set the Extended   
					Bolus duration in increments of 15 minutes for up   
					to 24 hours.   
					Note:   
					Press d to loop between programming the total bolus amount,   
					immediate bolus amount, and Extended Bolus duration.   
					7. Check the total and immediate bolus amounts and the extended bolus duration   
					programmed on the display. The set values can be viewed in the BOLUS   
					AMOUNT, IMMEDIATE BOLUS, and BOLUS DURATION screens.   
					8. Press f to confirm both bolus amounts and the bolus duration.   
					The MultiWave Bolus symbol ( ) blinks for 5 seconds (bolus delivery   
					start delay).   
					The pump beeps three times and the immediate bolus delivery begins. The   
					countdown of the remaining bolus continues to appear on the display.   
					Chapter 4: Boluses and TBRs   
					90   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					Once the immediate bolus is delivered, the remaining time, bolus amount,   
					and the current hourly basal rate are displayed on the RUN screen.   
					If a Temporary Basal Rate is active at the same time, the remaining time   
					and amount of the bolus and the amount of the increased or decreased   
					hourly basal rate are displayed on the RUN screen.   
					Cancelling a MultiWave Bolus   
					– during programming (a bolus amount or the bolus duration blinks):   
					There are three ways to cancel a MultiWave Bolus during programming:   
					• Do not press any key for 20 seconds. The pump returns to the RUN   
					screen (timeout).   
					• Exit to the MULTIWAVE BOLUS screen by pressing d+a simultaneously.   
					• Set the total bolus amount to 0.0 units and press f.   
					No bolus is delivered.   
					– during the start delay ( blinks):   
					• Press s or a. The pump beeps and vibrates.   
					The pump returns to the RUN screen.   
					An alert A8: BOLUS CANCELLED occurs. Press f twice to confirm   
					and turn off the alert.   
					No bolus is delivered.   
					See the section “Alert A8: BOLUS CANCELLED” (page 153) for more   
					information   
					– during bolus delivery:   
					• You can cancel the immediate delivery by pressing and holding s or   
					a for 3 seconds until you hear a melody. This cancels the whole   
					bolus (immediate and Extended Bolus delivery). An alert A8: BOLUS   
					CANCELLED occurs.   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					91   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					If the alert A8: BOLUS CANCELLED does not occur, the bolus delivery   
					was not cancelled. Cancel the extended delivery (see below).   
					• You can cancel the extended delivery by putting the pump in STOP   
					mode. This cancels the Extended Bolus. An alert A8: BOLUS CANCELLED   
					occurs. Press f twice to confirm and turn off the alert.   
					Bolus delivery is interrupted.   
					See the section “Alert A8: BOLUS CANCELLED” (page 153) for   
					more information.   
					You can review the actual bolus amount delivered before the cancellation   
					by looking at the bolus history. See the section “Review the Bolus History”   
					(page 107) for more information.   
					When your insulin pump is in STOP mode, it does not deliver   
					any insulin. Put your insulin pump in RUN mode to continue   
					the insulin delivery.   
					! 
					L 
					CAUTION   
					Note:   
					If a Temporary Basal Rate is active at the same time, it is also cancelled   
					when your pump is put in STOP mode. An alert A8: BOLUS CANCELLED   
					and an alert A6: TBR CANCELLED occur.   
					Press f twice to confirm and turn off the first alert. The second alert   
					appears on the display. Press f twice to confirm and turn off the second   
					alert. Both alerts are recorded in the alarm history.   
					See the sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8:   
					BOLUS CANCELLED” (page 153) for more information.   
					Make sure that the cancellation was intended and then program a new MultiWave   
					Bolus (and/or Temporary Basal Rate), if needed.   
					Chapter 4: Boluses and TBRs   
					92   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Programming a Temporary Basal Rate (TBR)   
					Your pump lets you temporarily increase or decrease your basal rate. This lets   
					you match changing insulin needs from increased or decreased activity level,   
					illness, or stress.   
					Incorrect programming of your insulin pump may cause   
					incorrect insulin delivery. Your insulin pump must be   
					programmed with your personal settings prior to starting   
					insulin pump therapy.   
					! 
					L 
					WARNING   
					Do not operate your insulin pump without knowing your   
					personal settings. If you are not sure about your personal   
					settings or if you are less experienced, have your doctor   
					or healthcare team check your personal settings.   
					Typically, a basal rate is set to 100%, but you can:   
					• Increase it to 200% for up to 24 hours   
					• Decrease it to 0% for up to 24 hours   
					When you increase or decrease your basal rate, each hourly basal rate for the   
					duration you program is increased or decreased. The characteristic shape of   
					your basal rate profile remains and does not become a flat curve, if the change   
					is larger than 0%. For 0% the basal rate profile becomes a flat curve and no   
					insulin (except boluses) is delivered during the programmed duration.   
					Discuss programming a Temporary Basal Rate with your doctor or healthcare team.   
					Note:   
					A running Temporary Basal Rate increase or decrease remains active,   
					even if you changed the basal rate profile.   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					93   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				RUN   
					screen   
					temporary   
					basal rate   
					precentage   
					duration   
					Follow the steps below:   
					1. Put your pump in RUN mode.   
					2. Press d to move to the TEMPORARY BASAL RATE (TBR) screen.   
					3. Press f to select.   
					The TBR PERCENTAGE screen appears.   
					Note:   
					If no Temporary Basal Rate is currently active, the percentage is set to   
					100%. If a Temporary Basal Rate is active, its duration and percentage   
					appear on the RUN screen.   
					4. Press a or s to increase or decrease the basal rate.   
					Chapter 4: Boluses and TBRs   
					94   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					When you increase (or decrease) the Temporary Basal Rate the duration of   
					your previous Temporary Basal Rate increase (or decrease) appears. When   
					you use your pump for the first time, a default value appears.   
					5. Press d to move to the TBR DURATION screen.   
					Press a or s to correct and set the duration.   
					Note:   
					Press d to loop between programming the basal rate percentage   
					and duration.   
					6. Press f to save and exit.   
					The new Temporary Basal Rate is immediately activated.   
					Note:   
					At any function screen you have three options to exit:   
					if you want to save the changes   
					• Press f. Your pump returns to the RUN screen.   
					if you want to undo the changes   
					• Wait for your pump to return to the RUN screen (timeout), or   
					• Press d+a simultaneously (exit feature) to exit the function screen.   
					• Set volume to 100% and press f.   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					95   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					During the Temporary Basal Rate delivery, the increase or decrease in   
					percentage, the increased or decreased hourly basal rate (hourly basal rate   
					plus TBR percentage), and the remaining time are displayed in the RUN   
					screen. An arrow pointing up symbolizes a Temporary Basal Rate increase;   
					an arrow pointing down symbolizes a Temporary Basal Rate decrease.   
					Note:   
					If an Extended Bolus or a MultiWave Bolus is active at the same time, the   
					remaining time and amount of the bolus and the amount of the increased   
					or decreased hourly basal rate are displayed in the RUN screen.   
					TBR OVER Alert   
					At the end of a Temporary Basal Rate, an alert A7: TBR OVER occurs. Press f   
					twice to confirm and turn off the alert. See the section “Alert A7: TBR OVER”   
					(page 153) for more information.   
					Chapter 4: Boluses and TBRs   
					96   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Cancelling a Temporary Basal Rate   
					– during programming:   
					There are three ways to cancel a Temporary Basal Rate during programming:   
					• Do not press any key for 20 seconds. The pump returns to the   
					RUN screen (timeout).   
					• Exit to the TEMPORARY BASAL RATE (TBR) screen by pressing d+a   
					simultaneously.   
					• Set the TBR percentage to 100%. Press f to save and exit.   
					No Temporary Basal Rate is delivered.   
					– during delivery:   
					There are two ways to cancel a Temporary Basal Rate during delivery:   
					• Set the TBR percentage to 100%:   
					1. Press d to move to the TEMPORARY BASAL RATE (TBR) screen.   
					2. Press f to select.   
					The TBR PERCENTAGE screen appears.   
					3. Press a or s to return the basal rate to 100%.   
					Press f to save and exit.   
					• Or, put your pump in STOP mode.   
					The Temporary Basal Rate delivery is interrupted.   
					An alert A6: TBR CANCELLED occurs. Press f twice to confirm and   
					turn off the alert.   
					Chapter 4: Boluses and TBRs   
					ACCU-CHEK Spirit Insulin Pump   
					97   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				When your insulin pump is in STOP mode, it does not deliver   
					any insulin. Put your insulin pump in RUN mode to continue   
					the insulin delivery.   
					! 
					L 
					CAUTION   
					Note:   
					If an Extended Bolus or a MultiWave Bolus is active at the same time,   
					it is also cancelled when your pump is put into STOP mode. An alert   
					A8: BOLUS CANCELLED and an alert A6: TBR CANCELLED occur.   
					Press f twice to confirm and turn off the first alert. The second alert   
					appears on the display. Press f twice to confirm and turn off the   
					second alert. Both alerts are recorded in the alarm history.   
					See the sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8:   
					BOLUS CANCELLED” (page 153) for more information.   
					Make sure that the cancellation was intended and then program a new   
					Temporary Basal Rate (and/or an Extended or MultiWave Bolus), if needed.   
					Chapter 4: Boluses and TBRs   
					98   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Chapter 5: User Menus   
					You will need to access various screens so you can view stored information,   
					program the pump, perform a function, or change a setting. Your pump offers   
					a choice of three different user menus:   
					• The STANDARD User Menu   
					This menu offers all the functions you need for successful insulin pump   
					therapy and lets you change between the user menus. If you are new   
					to pump therapy, start with this menu. Later, as your comfort with   
					pump therapy grows, you may choose the features offered in the   
					ADVANCED user menu.   
					• The ADVANCED User Menu   
					This menu offers the full range of the ACCU-CHEK Spirit insulin pump   
					functions. It includes all functions from the STANDARD user menu, plus   
					a wide range of functions for the more experienced user.   
					• The CUSTOM User Menu   
					Because the ACCU-CHEK Spirit insulin pump has many features, you   
					may want to choose the screens that you see when you scroll through   
					the menu choices. This menu lets you display or hide other screens.   
					Note:   
					At the end of this user guide you will find diagrams for the STANDARD   
					and ADVANCED user menus.   
					For your safety and convenience during programming, your insulin pump   
					returns itself to the RUN or STOP screen if no key is pressed within 20   
					seconds. For your safety, the changes you made before your insulin   
					pump “times out” will not be saved.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					99   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Selecting a User Menu   
					RUN   
					screen   
					select the   
					user menu   
					user   
					menu   
					select basal   
					rate profile   
					Note:   
					When you change the user menu, your current basal rate profile will not   
					appear if it is not activated. Make sure that your desired basal rate   
					profile number(s) is/are activated. This can be done through the previous   
					user menu or the ADVANCED user menu.   
					Follow the steps below:   
					1. Press d to move to the SELECT USER MENU screen.   
					Press f to select.   
					Chapter 5: User Menus   
					100   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				2. The current user menu is selected on the SELECT USER MENU screen.   
					Press a or s to select a different user menu.   
					3. Press f to confirm.   
					The selected basal rate profile and its daily insulin   
					total appear,   
					or   
					if the previous basal rate profile is not available   
					from the selected user menu, dashes appear instead.   
					4. Press a or s to select a basal rate profile, if needed.   
					Check the (new) basal rate profile and its daily insulin total.   
					5. Press f to save and exit. The new user menu and basal rate profile are   
					active immediately.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					101   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				A running increased or decreased Temporary Basal   
					Rate will remain active, even if you changed the basal   
					rate profile. The temporary basal rate is based on the   
					new basal rate.   
					! 
					L 
					CAUTION   
					Note:   
					At any function screen you have three options to exit:   
					• If you want to confirm and save the changes press f until your   
					pump returns to the RUN screen,   
					• If you want to undo the changes   
					press d+a simultaneously to exit the function screen, or   
					• Wait for your pump to return to the RUN screen (timeout).   
					Chapter 5: User Menus   
					102   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				STANDARD User Menu   
					The following is a list of menu choices for the STANDARD user menu as they   
					are addressed in the manual. To view the order in which they appear on the   
					pump, refer to the menu diagrams at the back of the manual. Some of the   
					functions have already been described earlier in the manual. The remaining   
					functions are described following this list.   
					SET TIME AND DATE–See the section “Setting Time and Date” (page 32)   
					PROGRAM BASAL RATE PROFILE - See the section “Programming your   
					Basal Rate Profile” (page 37)   
					CHANGE THE CARTRIDGE–See the section “Inserting the Cartridge” (page 49)   
					PRIME THE INFUSION SET–See the section “Priming the Infusion Set”   
					(page 53)   
					START YOUR ACCU-CHEK–See the section “Starting Insulin Delivery”   
					(page 65)   
					STOP YOUR ACCU-CHEK–See the section “Stopping Insulin Delivery”   
					(page 66)   
					STANDARD BOLUS–See the section “Programming a Standard Bolus”   
					(page 77)   
					EXTENDED BOLUS–See the section “Programming an Extended Bolus”   
					(page 83)   
					TEMPORARY BASAL RATE (TBR)–See the section “Programming a   
					Temporary Basal Rate (TBR)” (page 93)   
					COMMUNICATION–See the section “Data Transfer” (page 104)   
					INFORMATION–See the section “Reviewing Pump Data” (page 107)   
					SETUP MENU STANDARD–See the section “SETUP MENU Standard”   
					(page 113)   
					SELECT USER MENU–See the section “Selecting a User Menu” (page 100)   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					103   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Data Transfer   
					The pump memory stores up to 4,500 pump events (about 90 days of usage).   
					Pump events include alerts and errors, programming operations, and insulin   
					delivery records. Some or all of the pump memory may be transferable to   
					compatible software.   
					The built-in infrared interface inside the bottom cover of your insulin pump   
					allows for wireless data transfer.   
					Chapter 5: User Menus   
					104   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Setup for Data Transfer   
					STOP   
					screen   
					communi-   
					cation   
					start data   
					transfer   
					stop data transfer   
					Follow the steps below:   
					1. Put your pump in STOP mode.   
					2. Press d to move to the COMMUNICATION screen.   
					3. Press f. The DATA TRANSFER screen appears and a melody signals that   
					your pump is ready to deliver data.   
					When the data transfer is finished, press f to return to the STOP screen.   
					Set up your pump with a cartridge, adapter, and a new infusion set and put   
					it in the RUN mode, if necessary.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					105   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					Your pump returns to STOP mode 15 minutes after the last data transfer   
					when the DATA TRANSFER screen is selected.   
					You may press d+a simultaneously to exit while data is being   
					transferred. The data transfer is interrupted and an error E12: DATA   
					INTERRUPTED occurs. See the section, “Error E12: DATA INTERRUPTED”   
					(page 163) for more information.   
					Chapter 5: User Menus   
					106   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Reviewing Pump Data   
					The pump memory lets you review specific data such as:   
					• Bolus history   
					last 30 boluses   
					• Alarm history   
					last 30 alerts and errors   
					• History of daily insulin totals   
					• Temporary Basal Rate history   
					• Maintenance (pump timer)   
					last 30 daily totals of insulin delivered   
					last 30 increases and decreases   
					remaining time in days until the   
					pump timer expires   
					Review the Bolus History   
					The bolus history lets you review up to the last 30 boluses delivered beginning   
					with the most recent entry and working backward in time.   
					Each BOLUS HISTORY screen displays the:   
					• bolus type ( Standard Bolus, Extended Bolus or MultiWave Bolus),   
					• bolus duration ( only displayed if an Extended or a MultiWave Bolus),   
					• time (   
					) 
					• date ( ) and   
					• entry number (01 is the most recent) of total entries (example 01/30) of   
					a delivered bolus.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					107   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				RUN   
					screen   
					information   
					menu   
					bolus   
					history   
					alarm   
					his 
					daily   
					tot 
					TBR   
					his 
					1. Press d to move to the INFORMATION screen.   
					2. Press f to select.   
					The BOLUS HISTORY screen appears. The most   
					recent bolus information is displayed.   
					3. Press a or s to review the entries.   
					Press f to exit.   
					Chapter 5: User Menus   
					108   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Review the Alarm History   
					The alarm history lets you review the last 30 alerts or errors, beginning with   
					the most recent entry and working backward in time.   
					Each alarm history screen displays the:   
					• number and type of the alert or error (example A6),   
					• alert or error (example TBR CANCELLED),   
					• time ( ),   
					• date ( ) and   
					• entry number (01 is the most recent) of total entries (example: 01/30)   
					of an alert or error.   
					1. Press d to move to the INFORMATION screen.   
					Press f to select.   
					2. Press d to move to the ALARM HISTORY screen.   
					3. Press a or s to review the entries.   
					Press f to exit.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					109   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Review the History of Daily Insulin Totals   
					This history lets you review the last 30 daily totals of insulin delivered (from   
					midnight to midnight; basal rate plus boluses), starting with the most recent   
					entry and working backward in time.   
					Each daily insulin total screen displays the following history:   
					• daily total of insulin delivered ( ),   
					• its date ( ), and   
					• entry number (01 is the most recent) of total entries (example 01/30)   
					of a daily insulin total.   
					Follow the steps below:   
					1. Press d to move to the INFORMATION screen.   
					Press f to select.   
					2. Press d to move to the DAILY TOTALS screen.   
					3. Press a or s to review the entries.   
					Press f to exit.   
					Chapter 5: User Menus   
					110   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Review the Temporary Basal Rate History   
					The Temporary Basal Rate (TBR) history lets you review the last 30 TBR   
					increases or decreases starting with the most recent entry and working   
					backward in time.   
					Each TBR history screen displays the:   
					• TBR increase ( ) or decrease ( ) in percentage,   
					• TBR duration ( ),   
					• time (   
					) 
					• date ( ) and   
					• entry number (01 is the most recent) of total entries (example 01/30)   
					of a delivered Temporary Basal Rate.   
					Follow the steps below:   
					1. Press d to move to the INFORMATION screen.   
					Press f to select.   
					2. Press d to move to the TBR HISTORY screen.   
					3. Press or scroll a or s to review the entries.   
					Press f to exit.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					111   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Review the Time Remaining   
					Your pump is designed to be a highly reliable system that will give you worry-free   
					performance for a long time. For best performance, a pump timer limits the   
					operating time of the pump. This timer counts the number of days of operation   
					remaining for the ACCU-CHEK Spirit insulin pump.   
					Before the pump timer expires, an alert will remind you of the upcoming end of   
					operating time so that you can arrange to get a new one. The pump timer symbol   
					( 
					) appears in the RUN screen or STOP screen as a reminder. When the timer   
					expires, an error occurs and your pump goes into STOP mode. It can no longer   
					be put into RUN mode. See the sections “Alert A5: PUMP TIMER” (page 151)   
					and “Error E5: END OF OPERATION” (page 158) for more information.   
					Review the number of days left until the pump timer expires in the INFORMATION   
					menu.   
					Follow the steps below:   
					1. Press d to move to the INFORMATION screen.   
					Press f to select.   
					2. Press d to move to the PUMP TIMER screen.   
					The days until the pump timer expires is shown.   
					3. Press f to exit.   
					Chapter 5: User Menus   
					112   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				SETUP MENU Standard   
					The SETUP MENU Standard should not be confused with the STANDARD user   
					menu. The SETUP MENU Standard is a menu choice that appears in STANDARD   
					and ADVANCED user menus. The SETUP MENU Standard allows you to “set up”   
					certain features and settings. The following features and settings can be   
					changed using the SETUP MENU Standard menu choice.   
					Turning KeyLock On or Off   
					The KeyLock function lets you lock the four keys of your pump and serves as an   
					extra safety measure against activating functions by accident (for example, while   
					sleeping, during contact sports). To use the KeyLock function, it must be turned on.   
					See the section “Keys and Key Combinations” (page 11) for more information.   
					RUN / STOP   
					screen   
					setup   
					standard   
					KeyLock   
					beep   
					vo 
					alarm   
					sig 
					automatic   
					off   
					battery   
					typ 
					orientation   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					113   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Follow the steps below:   
					1. Press d to move to the SETUP MENU STANDARD screen.   
					Press f to select.   
					2. The current KeyLock status (ON or OFF) appears.   
					3. Press a or s to turn the KeyLock ON or OFF.   
					Press f to save and exit.   
					If no key is pressed for 10 seconds after your pump has returned to the RUN   
					screen or STOP screen and if the KeyLock is turned on, all four keys are locked.   
					Chapter 5: User Menus   
					114   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Unlocking the KeyLock   
					3 sec.   
					RUN screen   
					locked   
					10 sec.   
					RUN screen   
					unlocked   
					Follow the steps below:   
					1. From the RUN or the STOP screen, press d+s simultaneously (you hear   
					three beeps in RUN mode one beep in STOP mode) and hold for 3 seconds   
					until you again hear three beeps (in RUN mode) one beep (in STOP mode) to   
					confirm that the KeyLock is unlocked.   
					2. If no key is pressed for 10 seconds after your pump has returned to the RUN   
					or the STOP screen, all four keys are automatically locked.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					115   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Adjusting the Beep Volume   
					Your pump beeps when a key is pressed or when an alert or error occurs.   
					Follow the steps below:   
					1. Press d to move to the SETUP MENU STANDARD screen.   
					Press f to select.   
					2. Press d to move to the BEEP VOLUME screen.   
					The active beep volume is displayed.   
					3. Press a or s to adjust the beep volume.   
					There are five volume levels:   
					no beeps (beeps are turned off)   
					very low   
					low   
					normal   
					maximum   
					4. Press f to save and exit.   
					Note:   
					The volume of the STOP-Warning is independent of the programmed   
					beep volume. It always occurs at the maximum level. When the beeps   
					are turned off and an alert or error occurs, the beeps will become active   
					again after a short period of time to make sure that you are aware of   
					the alarm or error. These beeps will steadily increase in volume until   
					the maximum level is reached if the alert or error is not cancelled.   
					Chapter 5: User Menus   
					116   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Setting Up Alarm Signals   
					The ACCU-CHEK Spirit insulin pump lets you choose how you want your pump   
					to alert you when an alarm occurs.   
					You have three options:   
					beep only   
					vibrate only   
					beep and vibrate   
					Follow the steps below:   
					1. Press d to move to the SETUP MENU STANDARD screen.   
					Press f to select.   
					2. Press d to move to the ALARM SIGNALS screen.   
					The active alarm signals are displayed.   
					3. Press a or s to set the alarm signals.   
					4. Press f to save and exit.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					117   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Automatic Off   
					Discuss using the automatic off function with your doctor   
					or healthcare team.   
					! 
					L 
					CAUTION   
					Automatic off is a safety feature that stops insulin delivery and triggers an   
					error E3: AUTOMATIC OFF if no keys are pressed within a programmed time   
					period in RUN mode.   
					The automatic off can either be set to OFF or programmed up to 24 hours in   
					1-hour intervals in the SETUP MENU STANDARD screen.   
					Follow the steps below:   
					1. Press d to move to the SETUP MENU STANDARD screen.   
					Press f to select.   
					2. Press d to move to the AUTOMATIC OFF screen.   
					The current status is displayed at the bottom   
					left of the screen.   
					3. Press or scroll a to increase or s to decrease the duration for the   
					automatic off in 1-hour intervals.   
					4. If you would like to set the automatic off function to OFF, press or scroll s   
					until 0 hours and OFF appears on the display.   
					5. Press f to save and exit.   
					Chapter 5: User Menus   
					118   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Battery Type   
					If you do not set the correct battery type on your insulin   
					pump, the alert A2: BATTERY LOW may not occur in time   
					to provide you sufficient warning to replace your battery.   
					! 
					L 
					WARNING   
					Your pump lets you choose between using 1.5 volt AA alkaline batteries or   
					rechargeable AA NiMH batteries. If you change from one battery type to the   
					other, you must also change the battery type setting.   
					See the section “Accessories and Disposables–Battery” (page 21) for more   
					information on recommended batteries.   
					Follow the steps below:   
					1. Press d to move to the SETUP MENU STANDARD screen.   
					Press f to select.   
					2. Press d to move to the BATTERY TYPE screen.   
					The current battery type is displayed.   
					3. Press a or s to select the battery type ALKALINE or NiMH   
					(rechargeable) battery.   
					Press f to save and exit.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					119   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Display Orientation   
					Your pump lets you flip the display orientation by 180° so that you can view it   
					more conveniently depending on how you are wearing the pump.   
					Follow the steps below:   
					1. Press d to move to the SETUP MENU STANDARD screen.   
					Press f to select.   
					2. Press d to move to the ORIENTATION screen.   
					3. Press a or s to select the display orientation.   
					4. Press f to save and exit.   
					Chapter 5: User Menus   
					120   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				If you flip the display orientation of your insulin pump   
					by 180°, the a and s keys will also reverse their   
					functions. In comparison to the display orientation, the   
					upper key will be the a and the lower key will be s.   
					This change applies for all a and s functions including   
					turning on the backlight. The d and f will not change   
					in their function, but will remain the same, regardless of   
					your screen orientation.   
					! 
					L 
					CAUTION   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					121   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				ADVANCED User Menu   
					The ADVANCED user menu offers the full range of the ACCU-CHEK Spirit   
					insulin pump functions. The ADVANCED user menu includes all functions   
					from the STANDARD user menu, plus a wide range of functions for the more   
					experienced user.   
					The following is a list of additional menu choices that appear in the ADVANCED   
					user menu. Some of the functions have already been discussed earlier in the   
					manual. The remaining functions are described below.   
					EXTENDED BOLUS–See the section “Programming an Extended Bolus”   
					(page 83)   
					MULTIWAVE BOLUS–See the section “Programming a MultiWave Bolus”   
					(page 88)   
					CHANGE BASAL RATE PROFILE–See the section “Selecting a Basal   
					Rate Profile” (page 125)   
					PROGRAM BASAL RATE PROFILE–See the section “Programming   
					Additional Basal Sate Profiles” (page 124)   
					ALARM CLOCK–See the section “Alarm Clock” (page 127)   
					SETUP MENU ADVANCED–See the section “SETUP MENU Advanced”   
					(page 129)   
					Chapter 5: User Menus   
					122   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Basal Rate Profiles   
					Your pump can store up to five different basal rate profiles to match your   
					changing insulin needs (for example, Monday-Friday versus Exercise Day   
					versus Sleeping Late Day).   
					Discuss additional basal rate profiles with your doctor or healthcare team.   
					Consult with your doctor or healthcare team before changing basal rate profiles,   
					as this may change how you use your pump.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					123   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Programming Additional Basal Rate Profiles   
					Incorrect programming of your insulin pump may cause   
					! 
					incorrect insulin delivery. Your insulin pump must be   
					programmed with your personal settings prior to starting   
					insulin pump therapy. Do not operate your insulin pump   
					without knowing your personal settings. If you are not   
					sure about your personal settings or you are less   
					experienced, have your doctor or healthcare team   
					check your personal settings.   
					L 
					WARNING   
					Follow the steps below:   
					1. Press d to move to the PROGRAM BASAL RATE PROFILE screen.   
					Note:   
					Only basal rate profiles that are accessible in your current user menu   
					are available.   
					2. The hourly basal rates for each additional basal rate profile are set in the   
					same way as for basal rate profile 1. See the section “Programming your   
					Basal Rate Profile” (page 37) for more information.   
					Chapter 5: User Menus   
					124   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Selecting a Basal Rate Profile   
					RUN   
					screen   
					change the   
					BR profile   
					BR profile   
					no. …   
					Follow the steps below:   
					1. Press d to move to the CHANGE BASAL RATE PROFILE screen.   
					2. Press f to select.   
					The current basal rate profile is displayed.   
					Note:   
					Only basal rate profiles that are accessible in your current user menu   
					are available.   
					3. Press a or s to select a basal rate profile. The selected basal rate profile   
					and its daily insulin total appear.   
					4. Press f to select. The new basal rate profile is active immediately.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					125   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note:   
					When you change your user menu your current basal rate profile may   
					not appear.   
					If the basal rate profile number(s) you want does not appear,   
					• Select the ADVANCED user menu, or   
					• Program the basal rate profile to an available basal rate profile number   
					A running Temporary Basal Rate increase or decrease stays active,   
					even if you change the basal rate profile on your pump.   
					Chapter 5: User Menus   
					126   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Alarm Clock   
					You can set the alarm clock for a single alarm or for multiple alarms. Multiple   
					alarms repeat at the set time every day. Alarms can be set to remind you of   
					blood glucose testing or other important events.   
					Set Single or Multiple Alarms   
					1. Press d to move to the ALARM CLOCK screen.   
					2. Press f to select. The current alarm clock status (OFF, ONCE, or EVERY   
					DAY) and alarm time appear.   
					3. Press a or s to change the alarm clock to OFF, ONCE ( ), or EVERY   
					DAY ( ).   
					4. Press d to move to the SET ALARM HOUR screen.   
					Press a or s to set the hour.   
					5. Press d to move to the SET ALARM MINUTE screen.   
					Press a or s to set the minute.   
					6. Press f to save and exit.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					127   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Turning Off the Alarm Clock   
					When the alarm clock goes off, an alert A4: ALARM CLOCK occurs. Press f   
					twice to confirm and turn off the alert.   
					See the section “Alert A4: ALARM CLOCK” (page 151) for more information.   
					Chapter 5: User Menus   
					128   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				SETUP MENU Advanced   
					The SETUP MENU Advanced should not be confused with the ADVANCED user   
					menu. The SETUP MENU Advanced is a menu choice that appears in the   
					ADVANCED user menu. The SETUP MENU Advanced allows you to “set up” certain   
					features and settings. The following features and settings can be changed using   
					the SETUP MENU Advanced menu choice.   
					Time Format   
					You can set the time in either of the following formats:   
					American:   
					12-hour clock (AM/PM), example 1:39 PM   
					European/Military: 24-hour clock (00:00-23:59), example 13:39   
					Follow the steps below:   
					1. Press d to move to the SETUP MENU ADVANCED screen.   
					Press f to select.   
					2. The TIME FORMAT screen appears.   
					Press a or s to select a time format.   
					3. Press f to save and exit.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					129   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Date Format   
					You can set the date in the following formats:   
					American:   
					European:   
					mm/dd/yyyy, example 04/26/06   
					dd.mm.yyyy, example 26.04.06   
					Follow the steps below:   
					1. Press d to move to the SETUP MENU ADVANCED screen.   
					Press f to select.   
					2. Press d to move to the DATE FORMAT screen.   
					Press a or s to select a date format.   
					3. Press f to save and exit.   
					Chapter 5: User Menus   
					130   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Bolus increment   
					The bolus increment that can be programmed into your   
					! 
					insulin pump determines the “Quick” Standard Bolus   
					amount using the a and s keys of your insulin   
					pump. Incorrect programming of the bolus increment   
					may cause inappropriate insulin delivery.   
					L 
					WARNING   
					To address patients who may need large or small amounts of insulin, the bolus   
					increment for the ACCU-CHEK Spirit insulin pump is adjustable. The “Quick”   
					Standard Bolus is initially set to 0.5 units per key press but can be changed   
					in the SETUP MENU ADVANCED to 0.1, 0.2, 0.5, 1.0, or 2.0 units. Units for the   
					scroll “Standard” bolus, which may be adjusted by simply holding down the   
					a or s keys, will move at a constant 0.1 units. See the section   
					“Programming a Standard Bolus” (page 77) for more information.   
					Follow the steps below:   
					1. Press d to move to the SETUP MENU ADVANCED screen.   
					Press f to select.   
					2. Press d to move to the BOLUS INCREMENT screen.   
					Press a or s to select a bolus increment.   
					3. Press f to save and exit.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					131   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Prime Quantity   
					The prime quantity needed to fill an infusion set depends on the length of the   
					infusion set tubing being used. The shorter your infusion set tubing, the less   
					insulin will be required to prime the infusion set. The default setting for the   
					prime quantity is 25 units of U100 insulin.   
					Note:   
					The amount of insulin used for the priming is not added to the history of   
					daily insulin totals. You can press any of your pump’s keys to stop priming.   
					Follow the steps below:   
					1. Press d to move to the SETUP MENU ADVANCED screen.   
					Press f to select.   
					2. Press d to move to the PRIME QUANTITY screen.   
					Press a or s to select a prime quantity   
					between 0.0 and 30.0 units.   
					3. Press f to save and exit.   
					Chapter 5: User Menus   
					132   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Lock Basal Rate Profiles   
					Your pump lets you change your hourly basal rates as needed. You may lock   
					your basal rate profiles to provide extra protection against accidentally changing   
					a basal rate during normal operation. When this function is activated, your hourly   
					basal rates cannot be changed.   
					Follow the steps below:   
					1. Press d to move to the SETUP MENU ADVANCED screen.   
					Press f to select.   
					2. Press d to move to the BASAL RATE LOCK screen.   
					Press a or s to turn the basal rate lock ON   
					or OFF.   
					3. Press f to save and exit.   
					Note:   
					If the basal rate lock is set to ON (   
					), the programming of the basal   
					rate profiles 1, 2, 3, 4, and 5 is blocked. To check if the basal rate lock   
					is ON, press d to move to the PROGRAM BASAL RATE screen and   
					press f to select. The   
					indicates the lock of the basal rate programming. (The   
					symbol in the BASAL RATE TOTAL screen   
					symbol appears   
					in the BASAL RATE TOTAL screen only.) Press d to view the hourly   
					basal rates. The a and s functions are blocked.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					133   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Language   
					Your pump is programmed in several languages. These languages include   
					Italian, English, German, French, Spanish, Portuguese, Czech, Dutch, Swedish,   
					Norwegian, Polish, and Greek.   
					To choose a language, do the following:   
					1. Press d to move to SETUP MENU ADVANCED screen.   
					Press f to select.   
					2. Press d to move to the LANGUAGE screen.   
					Press a or s to select a language.   
					3. Press f to save and exit.   
					Chapter 5: User Menus   
					134   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Display Contrast   
					Your pump lets you adjust the display contrast.   
					Follow the steps below:   
					1. Press d to move to the SETUP MENU ADVANCED screen.   
					Press f to select.   
					2. Press d to move to the DISPLAY CONTRAST screen.   
					Press a or s to select a contrast level.   
					3. Press f to save and exit.   
					Chapter 5: User Menus   
					ACCU-CHEK Spirit Insulin Pump   
					135   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				CUSTOM User Menu   
					The CUSTOM user menu contains the 11 most commonly used functions according   
					to a customer survey.   
					The following is a list of the functions included in the CUSTOM user menu. All   
					of these functions have been addressed in detail in STANDARD and ADVANCED   
					user menu sections.   
					– STANDARD BOLUS   
					– EXTENDED BOLUS   
					– TEMPORARY BASAL RATE (TBR)   
					– INFORMATION   
					– CHANGE BASAL RATE PROFILE   
					– PROGRAM BASAL RATE PROFILE 1   
					– PROGRAM BASAL RATE PROFILE 2   
					– ALARM CLOCK   
					– SET DATE AND TIME   
					– SETUP MENU STANDARD   
					– SELECT USER MENU   
					Chapter 5: User Menus   
					136   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Chapter 6: Care and Maintenance   
					Care of Your ACCU-CHEK Spirit Insulin Pump   
					Do not perform any servicing or repair on your insulin   
					! 
					pump. For servicing or repairs contact Pump Support at   
					1-800-688-4578.   
					L 
					CAUTION   
					Disetronic Medical Systems, as the manufacturer, guarantees the functions of   
					your pump according to its specifications only if all servicing of your pump is   
					done and/or authorized by Disetronic Medical Systems.   
					System Inspection Checklist   
					Your pump must be well maintained to guarantee accurate insulin delivery.   
					Check the pump display at least every three hours during the day and before   
					you go to sleep, and especially if for any reason you might not be able to   
					hear the beeps or to feel the vibrations. Use only sterile products and   
					accessories designed to be used with your pump. Replace and discard these items   
					according to your doctor’s or healthcare team’s recommendations and the instruc-   
					tions that come with these items.   
					Chapter 6: Care and Maintenance   
					ACCU-CHEK Spirit Insulin Pump   
					137   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Check your pump daily. Make sure that:   
					• The pump casing, the display, and the cartridge are not chipped or   
					cracked and that the display doesn’t show incomplete or abnormal   
					letters or symbols.   
					• You visually inspect the cartridge. Make sure that the actual amount of   
					insulin in the cartridge equals the displayed cartridge amount on the   
					QUICK INFO screen.   
					• You inspect all parts of your pump and infusion set at least every   
					three hours during the day and before you go to sleep. If you detect   
					any loss of insulin, immediately replace the leaking component.   
					• The battery cover is correctly tightened and positioned even with the   
					pump casing.   
					• The adapter is inserted properly.   
					• Your infusion set is primed, free of air bubbles, and tight in the adapter.   
					• Your infusion set is inserted according to the instructions for its use.   
					• Your infusion site is secure, comfortable, and free of irritation or infection.   
					• The cartridge is free of air bubbles.   
					• Your pump is in RUN mode.   
					• The basal rates are set correctly according to your doctor’s or healthcare   
					team’s recommendations.   
					• The time and date are set correctly.   
					• The Temporary Basal Rate changes are set according to your doctor’s   
					or healthcare team’s recommendations.   
					• The beeps and/or vibrations are set.   
					• You have your personal emergency kit.   
					Chapter 6: Care and Maintenance   
					138   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				What to Do When You Drop Your Pump   
					If the pump is dropped it could be harmed or the seal that protects it against water   
					could be damaged. Be careful not to drop your pump. Use the proper ACCU-CHEK   
					Spirit insulin pump carrying system to help avoid dropping your pump. See   
					Disetronic Medical Systems sterile products and accessories catalogs and   
					brochures for more information. Or, contact Pump Support at 1-800-688-4578.   
					If your pump is dropped:   
					• Check that all connections in your infusion set are still tight, and if   
					they are not, reconnect or tighten them   
					• Check your pump, its sterile products, and accessories for chips and   
					cracks, and   
					• Change the cartridge if it is damaged.   
					Check your insulin pump and its sterile products and   
					accessories at least every three hours. Immediately inspect   
					your insulin pump if it is dropped. Do not use your insulin   
					pump if chips or cracks are visible. Chips and cracks may   
					allow water, dust, insulin, or other foreign substances to   
					enter your insulin pump and lead to malfunction.   
					! 
					L 
					CAUTION   
					If you have questions or are not sure whether your pump has been damaged,   
					contact Pump Support at 1-800-688-4578.   
					Chapter 6: Care and Maintenance   
					ACCU-CHEK Spirit Insulin Pump   
					139   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Storing Your ACCU-CHEK Spirit Insulin Pump   
					When your insulin pump will not be in use for a long period of   
					time, it should be properly stored, to prevent a malfunction.   
					! 
					L 
					CAUTION   
					To store your pump, place in STOP mode, then:   
					• Remove the battery to preserve the battery life   
					• Remove the cartridge   
					• Insert the battery cover and the adapter   
					• Store your pump in its shipping case   
					Storage Conditions   
					Temperature   
					Air humidity   
					Barometric pressure   
					41°F to 113°F (+5°C to +45°C)   
					5 to 85% relative air humidity   
					700 to 1060 mbar   
					Disposal   
					Your ACCU-CHEK Spirit Insulin Pump   
					If you need to dispose of your insulin pump, return it to Disetronic Medical   
					Systems for professional disposal.   
					Battery   
					Properly dispose of dead batteries.   
					Cartridge, Infusion Set, Adapter, and Accessories   
					Dispose of these items in your normal garbage. Remember to dispose of the   
					introducer needle safely (needle container). To prevent injury to others, replace   
					the needle protective cap or insert the needle of the infusion set into a disposable   
					object, or put it in a hard shell container.   
					Chapter 6: Care and Maintenance   
					140   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Cleaning Your ACCU-CHEK Spirit Insulin Pump   
					The best time to clean your pump is when you change the cartridge. Use a   
					soft, dry cloth for cleaning. If the pump compartments are very dirty, contact   
					Pump Support at 1-800-688-4578 for more instructions.   
					Always remove the cartridge and adapter and place your   
					insulin pump in STOP mode while cleaning. Avoid pressing   
					! 
					L 
					the keys of your insulin pump during cleaning, as this may   
					accidentally change your settings. Do not use alcohol,   
					solvents, strong detergents, bleaching agents, scouring   
					pads, or sharp instruments for cleaning as they may   
					damage the pump.   
					CAUTION   
					Battery Information   
					It is important to change your battery regularly and to have spare batteries ready.   
					The life of any battery is affected by how you use your insulin pump, your personal   
					settings, delivery rates, temperature, and other factors. With typical pump use   
					[50 U/day using U100 insulin; operating temperature 72°F 6°F (22°C 3°C)],   
					pump battery life is about four weeks using alkaline batteries and one week   
					using rechargeable batteries. See the section “Accessories and Disposables–   
					Battery” (page 21) for more information on recommended batteries. See the   
					section “Inserting (or Changing) the Battery” (page 25) for instructions on how to   
					change your battery.   
					Chapter 6: Care and Maintenance   
					ACCU-CHEK Spirit Insulin Pump   
					141   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Repair   
					Do not perform any servicing or repair on your insulin pump.   
					Do not use any lubricants on the pump mechanism. If you   
					have questions, contact Pump Support at 1-800-688-4578.   
					! 
					L 
					CAUTION   
					If you have problems with your pump, contact Pump Support at 1-800-688-4578.   
					Pump Support can help you and answer many questions over the telephone.   
					Pump Support may direct you to return your pump to Disetronic Medical Systems   
					for repair. For example, you might need to return your pump if an alert or error   
					could not be cleared by following the procedures described in “Alerts and Errors.”   
					(page 147). If Pump Support directs you to return your pump, they will give you   
					a Return Material Authorization (RMA) number that you must include with the   
					pump when you send it.   
					If you must ship your pump, carefully pack the pump and the used cartridge,   
					battery, battery cover, adapter, and infusion set to avoid damage to them during   
					transportation. Place your pump back in the shipping case and put the case in a   
					package that will not rip or tear. It is best to send your pump using a shipping   
					method and carrier that tracks the shipment.   
					When you return your pump, include the following:   
					• a description of the reason for returning your pump   
					• your name and address   
					• your daytime phone   
					• the serial number of your pump, and   
					• any RMA (Return Material Authorization) number received from   
					Pump Support.   
					By returning your pump to Disetronic Medical Systems, you allow Disetronic   
					Medical Systems to undertake any testing on your infusion pump that is needed for   
					a proper inspection, and understand that your pump may not be returned to you.   
					Chapter 6: Care and Maintenance   
					142   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Troubleshooting   
					This section describes problems that may occur while using your ACCU-CHEK   
					Spirit insulin pump. See “Alerts and Errors” (page 147) for more information.   
					Note:   
					The solutions in this guide are only suggestions. Do not use these   
					procedures without approval of your doctor or healthcare team. Always   
					follow your doctor’s or healthcare team’s recommendations, and   
					contact them if you have any concerns about your insulin pump therapy.   
					You need to know that the interruption of insulin delivery (due to a leak,   
					occlusion, or loss of insulin potency, for example) or the malfunction of   
					your insulin pump can result in a rapid rise of your blood glucose level.   
					Although your pump has an internal security check system, it cannot   
					alert you if your infusion set is leaking or if the insulin you are using   
					has lost its potency.   
					Chapter 6: Care and Maintenance   
					ACCU-CHEK Spirit Insulin Pump   
					143   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Therapy-Related Issues   
					Many factors, such as alcohol consumption, medications other than insulin,   
					ineffective or expired insulin, decreased activity, illness, and stress can affect   
					blood glucose levels. Contact your doctor or healthcare team for advice on how   
					to manage these or any other therapy related issues.   
					Troubleshooting Table   
					Problem   
					Recommended Actions   
					Individual characters, numbers, or Contact Pump Support at   
					symbols on the display are shown 1-800-688-4578.   
					incompletely or not at all.   
					Your insulin pump does not beep Check that the beep volume and alarm   
					or vibrate.   
					signals are turned on. Check and   
					change the battery. If your insulin pump   
					still does not beep or vibrate, contact   
					Pump Support at 1-800-688-4578.   
					Self-test incomplete (after an error Remove the battery, wait a few   
					E7: ELECTRONIC ERROR).   
					seconds, and insert it again. If the   
					self-test remains incomplete, contact   
					Pump Support at 1-800-688-4578.   
					Incorrect insulin delivery due to   
					incorrect time setting.   
					The correct hourly basal rate cannot be   
					delivered if the time is not set correctly.   
					Check and reset the time if it is wrong.   
					Do the same check when you change   
					the battery.   
					Incorrect insulin delivery due to   
					Always review changes made in basal   
					incorrect basal rate programming. rate programming. Confirm that you are   
					using your correct basal rate profile.   
					Chapter 6: Care and Maintenance   
					144   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Problem   
					Recommended Actions   
					Air bubbles in the cartridge   
					and/or infusion set   
					Check blood glucose level. In the   
					cartridge: Disconnect the infusion tube   
					set from the headset and re-prime your   
					infusion tube set. If the air bubbles   
					remain, replace the cartridge. Make sure   
					to fill the cartridge with insulin at room   
					temperature. In the infusion set:   
					Disconnect and prime your infusion set.   
					Reconnect and make sure all connections   
					are tight and secure.   
					Blood in the infusion set tubing Change your infusion set and infusion site.   
					Empty cartridge   
					Replace the cartridge. Cartridges are for   
					single use only.   
					Disconnected or dislodged   
					infusion set   
					Make sure all connections are tight. Inspect   
					connections for leaking insulin. Check your   
					blood glucose level. Change your infusion   
					set and infusion site.   
					Leak in system   
					Check for leaking insulin at all connections   
					and on the skin. Change your infusion set,   
					infusion site, cartridge, and adapter. Check   
					your blood glucose level.   
					Chapter 6: Care and Maintenance   
					ACCU-CHEK Spirit Insulin Pump   
					145   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Problem   
					Recommended Actions   
					Occlusion in system   
					Disconnect and re-prime your infusion   
					set. Check your blood glucose. If the   
					occlusion remains, prime the pump   
					without the cartridge inserted. See the   
					section “Error E4: OCCLUSION” (page 156).   
					Failure to complete prime   
					Poor absorption of insulin   
					Re-prime the infusion set. Always prime   
					until a bubble-free insulin flow emerges   
					from the tip of the needle. Make sure the   
					tubing is free of air.   
					Choose infusion sites that do not have   
					scar tissue, bruises, or tissue build-up.   
					Do not use the same infusion set or   
					infusion site longer than your doctor or   
					healthcare team recommends.   
					Infusion site becomes sore,   
					red, or swollen.   
					Change your infusion headset and infusion   
					site immediately. Use proper insertion   
					technique. Check blood glucose level.   
					Follow the planned infusion site rotation   
					pattern and advice given by your doctor or   
					healthcare team.   
					Chapter 6: Care and Maintenance   
					146   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Chapter 7: Alerts and Errors   
					A safety system monitors and controls the function of your pump. This safety   
					system performs more than 9 million safety checks every day. If it detects a   
					change from the pump’s normal state, it will trigger an alert (warning instruction)   
					or an error (error message). Beeps, vibrations, and messages on the pump   
					display indicate the cause of an alert or error. Alerts and errors are the only   
					way you will be quickly notified of any changes in your pump.   
					Check the pump display at least every three hours during   
					the day, and before you go to sleep, and especially if you   
					might be not be able to hear the beeps or feel the vibrations.   
					! 
					L 
					CAUTION   
					Your pump gives alerts and errors using beeps and vibrations. You can turn off   
					either the beeps or the vibrations, but for your safety you cannot turn off both   
					at the same time for the initial alarm signal.   
					If you do not confirm an alert or error within 60 seconds, it will signal again.   
					When the beep is turned off or programmed at lower than maximum volume   
					level, the beep volume will rise to the loudest volume level if not confirmed.   
					The beep and vibration will continue at this level until confirmed.   
					Note:   
					The volume of the STOP-Warning is separate from the programmed beep   
					volume and cannot be changed. It always occurs at the maximum level.   
					KeyLock is unlocked by the system each time an alert or error occurs.   
					Chapter 7: Alerts and Errors   
					ACCU-CHEK Spirit Insulin Pump   
					147   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Confirming an Alert or Error   
					When you confirm A1: CARTRIDGE LOW   
					, E1: CARTRIDGE EMPTY,   
					A2: BATTERY LOW , E2: BATTERY DEPLETED, A5: PUMP TIMER   
					or E5:   
					END OF OPERATION, the symbol for that alert or error stays in the RUN screen   
					or STOP screen as a reminder. For other alerts or errors, alarm information is   
					cleared from the display but stays in the alarm history and the event memory.   
					Follow the steps below to confirm an alert or error:   
					1. Press f to turn off the beeps and vibrations.   
					The alert or error code stays on the display.   
					2. As soon as you understand the reason for the alert, error code, or message,   
					confirm by pressing f again.   
					3. Take the appropriate action, if any.   
					4. If more than one alert and/or error occurs simultaneously, press f twice   
					for each alert or error to confirm and turn each off.   
					5. After an alert or error, make sure that your pump is returned to RUN mode.   
					Chapter 7: Alerts and Errors   
					148   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				List of Alerts   
					After an alert occurs, your insulin pump may be in STOP   
					! 
					mode and the insulin delivery may be interrupted. In order   
					to maintain insulin delivery, you must act immediately   
					according to the instructions given for each error code   
					or put your insulin pump in RUN mode to continue the   
					insulin delivery.   
					L 
					WARNING   
					Alert A1: CARTRIDGE LOW   
					The contents of the insulin cartridge have dropped to 20 units of insulin. Be   
					prepared to replace the cartridge as soon as possible. The cartridge low symbol   
					( 
					) appears in the RUN or STOP screen as a reminder until you change   
					the cartridge.   
					1. Press f twice to turn off and to confirm the alert.   
					Change the cartridge before it is empty.   
					2. If necessary, put your pump in RUN mode.   
					See the section “Changing the Cartridge and Infusion Set” (page 73) for more   
					information.   
					Chapter 7: Alerts and Errors   
					ACCU-CHEK Spirit Insulin Pump   
					149   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Alert A2: BATTERY LOW   
					The battery power is low. Be prepared for a battery change as quickly as possible.   
					The battery low symbol (   
					) appears in the RUN or STOP screen as a reminder   
					until you change the battery.   
					1. Press f twice to turn off and to confirm the alert.   
					Replace the battery as soon as possible (within a   
					few hours). Failure to change the battery within a   
					few hours of receiving the alert could result in   
					additional error messages.   
					2. If necessary, put your pump in RUN mode.   
					See the section “Accessories and Disposables–Battery” (page 21) for more   
					information on recommended batteries. See the section “Inserting (or Changing)   
					the Battery” (page 25) for instructions on how to change your battery.   
					Alert A3: REVIEW TIME AND DATE   
					The pump was without power for more than one hour. The ACCU-CHEK Spirit   
					insulin pump settings and history data are saved in memory regardless of   
					power loss. However, it may be necessary to enter the time and date again.   
					1. Press f twice to turn off and to confirm the alert.   
					Review the time and date and correct them,   
					if they are wrong.   
					2. If necessary, put your pump in RUN mode.   
					See the section “Setting Time and Date” (page 32) for more information.   
					Chapter 7: Alerts and Errors   
					150   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Alert A4: ALARM CLOCK   
					An alarm clock reminder was programmed into the pump.   
					1. Press f twice to turn off and to confirm the alert.   
					2. If necessary, put your pump in RUN mode.   
					Note:   
					If you have set the alarm occurrence to EVERY DAY, the same alert occurs   
					every 24 hours. Turn the alarm clock function OFF if you do not want   
					the alarm to go off again at that same time every day.   
					See the section “Alarm Clock” (page 127) for more information.   
					Alert A5: PUMP TIMER   
					Your pump’s operating time will expire soon. It is recommended that you replace   
					your pump after a period of four years of running time. After the alert A5: PUMP   
					TIMER occurs, the ACCU-CHEK Spirit insulin pump will display the symbol   
					remind you to check the timer.   
					to   
					1. Press f twice to turn off and to confirm the alert.   
					2. Put your pump in RUN mode.   
					3. Go to the INFORMATION screen on your pump.   
					Chapter 7: Alerts and Errors   
					ACCU-CHEK Spirit Insulin Pump   
					151   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				4. Move to the PUMP TIMER screen.   
					The days until the pump timer expires are shown.   
					When the PUMP TIMER reaches zero, the pump will change to STOP mode and   
					no longer operate. An error, E5: END OF OPERATION, will appear on the display.   
					Contact Pump Support at 1-800-688-4578 to discuss your options for further   
					use of your ACCU-CHEK Spirit insulin pump.   
					Note:   
					The alert A5: PUMP TIMER ALERT and the error E5: END OF OPERATION   
					are based on running time (the total time the pump is in RUN mode)   
					and not calendar years. However, the warranty is based on the date the   
					pump was purchased.   
					See the section “Review the Time Remaining” (page 112) for more information.   
					Alert A6: TBR CANCELLED (Temporary Basal Rate Cancelled)   
					A temporary increase or decrease of the basal rate has been cancelled.   
					1. Press f twice to turn off and to confirm the alert.   
					2. If necessary, put your pump in RUN mode.   
					3. Make sure that the cancellation was intended and then program a new   
					Temporary Basal Rate if needed.   
					Chapter 7: Alerts and Errors   
					152   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Alert A7: TBR OVER (Temporary Basal Rate Over)   
					A Temporary Basal Rate has ended. The original basal rate (100%) will resume   
					automatically once the alert has been confirmed.   
					1. Press f twice to turn off and confirm the alert.   
					2. Decide whether a further Temporary Basal Rate change is needed. If necessary,   
					program a new Temporary Basal Rate.   
					Alert A8: BOLUS CANCELLED   
					A bolus was cancelled during the start delay or after delivery began.   
					1. Press f twice to turn off and to confirm the alert.   
					2. If necessary, put your pump in RUN mode.   
					3. Make sure that the cancellation was intended and then program a new   
					bolus if needed.   
					Note:   
					The actual bolus amount that was delivered before the cancellation can be   
					reviewed in the bolus history. See the section “Review the Bolus   
					History” (page 107) for more information.   
					Chapter 7: Alerts and Errors   
					ACCU-CHEK Spirit Insulin Pump   
					153   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				List of Errors   
					After the occurrence of an error, your insulin pump will be in   
					! 
					STOP mode and the insulin delivery is interrupted. In order   
					to maintain insulin delivery, you must act immediately   
					according to the instructions given for each error code.   
					L 
					WARNING   
					Error E1: CARTRIDGE EMPTY   
					The insulin cartridge is empty. Change the cartridge immediately.   
					1. Press f twice to turn off and to confirm the error.   
					2. Disconnect or remove the infusion set from your infusion site.   
					3. Change the cartridge and prime the infusion set.   
					4. Reconnect the infusion set at your infusion site.   
					5. Put your pump in RUN mode.   
					See the section “Changing the Cartridge and Infusion Set” (page 73) for more   
					information.   
					Chapter 7: Alerts and Errors   
					154   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Error E2: BATTERY DEPLETED   
					The battery is out of power. Replace it immediately.   
					After inserting a new battery, always check that the time   
					! 
					and date of your insulin pump are set correctly. Incorrect   
					programming of the time and date may cause incorrect   
					insulin delivery. If you, your doctor, or healthcare team   
					store and analyze your therapy data electronically and the   
					time and date of the devices used are not set identically,   
					the gathered data might not be meaningful.   
					L 
					WARNING   
					1. Press f twice to turn off and to confirm the error.   
					2. Replace the battery.   
					3. Put your pump in RUN mode.   
					Note:   
					Your pump stores the settings you saved before the battery power ran   
					out. If your pump is without power for more than one hour, check the   
					time and date.   
					See the section “Accessories and Disposables–Battery” (page 21) for more   
					information on recommended batteries. See the section “Inserting (or changing)   
					the Battery” (page 25) for instructions on how to change your battery.   
					Chapter 7: Alerts and Errors   
					ACCU-CHEK Spirit Insulin Pump   
					155   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Error E3: AUTOMATIC OFF   
					Automatic off is a safety feature that stops insulin delivery and triggers an   
					error E3: AUTOMATIC OFF if no keys are pressed within a programmed time   
					period in RUN mode.   
					1. Press f twice to turn off and to confirm the error.   
					2. Put your pump in RUN mode.   
					Error E4: OCCLUSION   
					If an error E4: OCCLUSION occurs, immediately check your   
					blood glucose level, and check again in about 2 hours,   
					because insulin delivery has been interrupted. If your blood   
					glucose level is high, take appropriate actions according to   
					your doctor’s or healthcare team’s instructions.   
					! 
					L 
					WARNING   
					Insulin (a maximum of 3.5 units of U100 insulin) was not delivered. This may   
					be due to:   
					• a blocked infusion set,   
					• a reused cartridge, or   
					• a dirty or damaged piston rod.   
					Note:   
					The display of this error depends on several factors including your current   
					hourly basal rate and boluses. For example, a blocked infusion set with   
					a basal rate of 2.0 U/h (U100 insulin and no additional boluses) would   
					trigger an error E4: OCCLUSION in approximately three hours or less,   
					depending on the type and severity of the blockage.   
					Chapter 7: Alerts and Errors   
					156   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				If the occlusion (blockage) was inside your infusion set, follow the   
					steps below.   
					1. Press f twice to turn off and to confirm the error.   
					2. Disconnect or remove the infusion set from your infusion site.   
					3. Change the infusion set and prime it.   
					4. Reconnect the infusion set (tubing) at your infusion site.   
					5. Put your pump in RUN mode.   
					In rare cases, the cartridge itself can be the cause of the occlusion. If an   
					error E4: OCCLUSION occurs again, you must change the cartridge. Follow   
					the steps below:   
					1. Press f twice to turn off and to confirm the error.   
					2. Disconnect or remove the infusion set from your infusion site.   
					3. Change the cartridge and prime the infusion set.   
					4. Reconnect the infusion set (tubing) at your infusion site.   
					5. Put your pump in RUN mode.   
					Chapter 7: Alerts and Errors   
					ACCU-CHEK Spirit Insulin Pump   
					157   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				If the error E4: OCCLUSION occurs again, prime your pump without a   
					cartridge. Follow the steps below:   
					1. Press f twice to turn off and to confirm the error.   
					2. Disconnect or remove the infusion set from your infusion site.   
					3. Remove the cartridge, adapter and infusion set from your pump.   
					4. Carry out the PRIME THE INFUSION SET function without the cartridge,   
					adapter, and infusion set inserted.   
					5. Insert a new cartridge with connected adapter and infusion set.   
					6. Prime the infusion set.   
					7. Reconnect the infusion set (tubing) at your infusion site.   
					8. Put your pump in RUN mode.   
					If the error E4: OCCLUSION continues to occur, contact your doctor or   
					healthcare team for an alternate therapy plan. Contact Pump Support at   
					1-800-688-4578 for more assistance.   
					Error E5: END OF OPERATION   
					Your pump timer has reached zero. The pump will change to STOP mode and will   
					no longer operate. An error E5: END OF OPERATION will appear on the display. It   
					is recommended that you replace your pump after four years. Before this error   
					occurs, an alert A5: PUMP TIMER will remind you to replace your pump soon.   
					1. Press f twice to turn off and to confirm the error.   
					Chapter 7: Alerts and Errors   
					158   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				2. Contact Pump Support at 1-800-688-4578 to discuss your options for further   
					use of the ACCU-CHEK Spirit insulin pump.   
					Note:   
					The alert A5: PUMP TIMER ALERT and the error E5: END OF OPERATION   
					are based on running time (the total time the pump is in RUN mode)   
					and not calendar years. However, the warranty is based on the date the   
					pump was purchased.   
					Error E6: MECHANICAL ERROR   
					A mechanical error was discovered during a routine automatic system check   
					and insulin delivery has stopped.   
					1. Press f twice to turn off and to confirm the error.   
					2. Remove or disconnect the infusion set from your infusion site.   
					3. Remove the cartridge, adapter, infusion set (tubing), and battery from   
					your pump. Insert a new battery.   
					4. Rewind the piston rod.   
					5. Install a new cartridge and infusion set.   
					6. Prime the infusion set.   
					7. Reconnect the infusion set (tubing) at your infusion site.   
					8. Put your pump in RUN mode.   
					Chapter 7: Alerts and Errors   
					ACCU-CHEK Spirit Insulin Pump   
					159   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				If the error E6: MECHANICAL ERROR occurs again in spite of these actions,   
					contact your doctor or healthcare team for an alternate therapy plan. Contact   
					Pump Support at 1-800-688-4578 for more assistance.   
					Error E7: ELECTRONIC ERROR   
					An electronic error was discovered during a routine automatic system check   
					and the insulin delivery has stopped. You cannot turn off and confirm an error   
					E7: ELECTRONIC ERROR by pressing f. You must remove the battery instead.   
					1. Remove or disconnect the infusion set from your infusion site.   
					2. Insert a new battery into the pump.   
					3. Prime the infusion set.   
					4. Reconnect the infusion set (tubing) at your infusion site.   
					5. Put your pump in RUN mode.   
					If an error E7: ELECTRONIC ERROR occurs during programming (i.e., basal rate,   
					time and date) check your settings.   
					If the error E7: ELECTRONIC ERROR occurs again, contact Pump Support at   
					1-800-688-4578 for assistance. Contact your doctor or healthcare team if   
					you need an alternate therapy plan.   
					Chapter 7: Alerts and Errors   
					160   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Error E8: POWER INTERRUPT   
					A power interrupt error occurs when you insert a new battery without first placing   
					the pump in STOP mode. See the section “Inserting (or Changing) the Battery”   
					(page 25) for instructions on how to change your battery. The error can also occur   
					if the pump was dropped. If the pump was dropped, see the section “What to Do   
					When You Drop Your Pump” (page 139) for more information.   
					1. Press f twice to turn off and to confirm the error.   
					2. Your pump goes in STOP mode.   
					3. Check the time and date and correct them if they are wrong. See the section   
					“Setting Time and Date” (page 32) for more information.   
					4. If a bolus and/or a Temporary Basal Rate was interrupted by the error,   
					review the bolus and/or the Temporary Basal Rate history for the delivered   
					amount and duration. See the section “Reviewing Pump Data” (page 107)   
					for instructions.   
					5. Put your pump in RUN mode.   
					6. Program a new bolus and/or Temporary Basal Rate if you need them. See   
					“Boluses and Temporary Basal Rates” (page 75).   
					If an error E8: POWER INTERRUPT occurs during programming, check your settings.   
					Chapter 7: Alerts and Errors   
					ACCU-CHEK Spirit Insulin Pump   
					161   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Error E10: CARTRIDGE ERROR   
					The CHANGE THE CARTRIDGE function was not correctly carried out.   
					1. Press f twice to turn off and to confirm the error.   
					2. Remove or disconnect the infusion set from your infusion site.   
					3. Start the CHANGE THE CARTRIDGE menu.   
					4. Prime the infusion set.   
					5. Reconnect the infusion set (tubing) at your infusion site.   
					6. Put your pump into RUN mode.   
					7. If an error E10: CARTRIDGE ERROR occurs during the winding of the piston   
					rod, insert a new battery into the pump and repeat steps 1-6.   
					Error E11: SET NOT PRIMED   
					Never prime an infusion set that is connected to your body.   
					You risk uncontrolled insulin delivery into your body. With   
					detachable infusion sets, make sure you have disconnected   
					the tubing from your site prior to priming. Always follow   
					the instructions provided with your infusion set.   
					! 
					L 
					WARNING   
					The cartridge was changed but the infusion set was not primed. The error will   
					display when you try to put your pump in RUN mode.   
					Chapter 7: Alerts and Errors   
					162   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				1. Press f twice to turn off and to confirm the error.   
					2. Remove or disconnect the infusion set from your infusion site.   
					3. Prime the infusion set.   
					4. Reconnect the infusion set (tubing) at your infusion site.   
					5. Put your pump in RUN mode.   
					See the sections “Changing the Cartridge and Infusion Set” (page 73) and   
					“Priming the Infusion Set” (page 53) for more information.   
					Error E12: DATA INTERRUPTED   
					If data transfer between compatible software and your   
					insulin pump is disrupted in any way, the configuration   
					may be incomplete and an error E12: DATA INTERRUPTED   
					occurs. The data transfer must be completed successfully   
					before you can put your insulin pump into RUN mode.   
					! 
					L 
					CAUTION   
					Data transfer between your pump and a PC was interrupted before it was   
					completed.   
					1. Press f twice to turn off and to confirm the error.   
					2. Restart the data transfer.   
					Before you can put your pump in RUN mode, the data transfer must be successful.   
					Chapter 7: Alerts and Errors   
					ACCU-CHEK Spirit Insulin Pump   
					163   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				See the section “Data Transfer” (page 104) for more information.   
					Error E13: LANGUAGE ERROR   
					The language setting needs to be checked.   
					1. Press f twice to turn off and to confirm the error.   
					2. Follow the steps in the section “SETUP MENU Advanced–Language”   
					(page 134) to change the language.   
					3. Put your pump in RUN mode.   
					Chapter 7: Alerts and Errors   
					164   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Appendixes   
					Appendix A: Technical Data   
					General Technical Data   
					Air humidity   
					• During operation: 20 to 90%   
					relative humidity   
					• Stored in its shipping case:   
					5 to 85% relative humidity   
					Barometric pressure   
					During operation or stored in its   
					shipping case: 700 to 1,060   
					mbar [do not exceed 10,000 feet   
					(3,000 meters) above sea level].   
					Basal rate   
					Battery life   
					Minimum 0.1 U/h; maximum 25 U/h   
					There are 24 hourly basal rates   
					adjustable in 0.1 unit increments.   
					If used in a typical usage pattern   
					[50 U/day using U100 insulin;   
					operating temp. of 72°F 6°F   
					(22°C 3°C)] the battery life is   
					approx. four weeks for alkaline   
					and one week for rechargeable   
					batteries.   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					165   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Bolus   
					The max. bolus amount per delivery   
					is 25 units when using U100 insulin.   
					The bolus amount for the “Quick”   
					Standard Bolus is adjustable in   
					increments of 0.1, 0.2, 0.5, 1.0,   
					and 2.0 units. For the “Scroll”   
					Standard Bolus, the Extended   
					Bolus, and the MultiWave Bolus,   
					the amount is adjustable in fixed   
					increments of 0.1 units. The   
					duration of the Extended Bolus and   
					the MultiWave Bolus is adjustable in   
					intervals of 15 minutes (15 minutes   
					up to 24 hours).   
					Cartridge   
					ACCU-CHEK 3.15 ml plastic   
					cartridge with luer-lock connection.   
					Data storage time   
					The time and date is safely stored   
					in the memory for about 1 hour   
					after the battery has been removed.   
					Your insulin pump’s settings (the   
					hourly basal rates, remaining   
					cartridge content, bolus increments,   
					and active user menu) and the event   
					memory (bolus history, history of   
					daily insulin totals, Temporary   
					Basal Rate history, alarm history)   
					are saved, regardless of battery   
					voltage and time the insulin pump   
					has been without a battery.   
					Data transfer   
					Delivery   
					Infrared interface   
					1/20th of the current hourly basal   
					rate in 3-minute intervals.   
					Dimensions (maximum)   
					without adapter   
					Approx. 3.2 x 2.2 x 0.8 inches   
					(81 x 55 x 20 mm)   
					Appendixes   
					166   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Flow (delivery rate)   
					0.2 U/sec during infusion set   
					priming and bolus when using   
					U100 insulin.   
					Infusion sets   
					IPX 8   
					ACCU-CHEK infusion sets with   
					luer-lock connector.   
					Protected against the effects of   
					temporary immersion in water   
					under standardized conditions   
					[up to 60 minutes and 8 feet   
					(2.5 meters)].   
					Maximum time for an error   
					E4: OCCLUSION   
					Plastic cartridges*   
					• at med. basal rate 1.0 U/h using   
					U100 insulin   
					≤ 5h   
					• at min. basal rate 0.1 U/h using   
					U100 insulin   
					U100 insulin   
					≤ 50h   
					≤ 3.5 U   
					Maximum volume* before error   
					Maximum pressure   
					400kPa (4.0 bar)   
					U100 insulin   
					Maximum quantity delivered at   
					a single fault condition   
					≤ 1.0 U   
					Power supply   
					One 1.5 volt AA Alkaline battery   
					or one rechargeable Nickel Metal   
					Hydride (NiMH) AA battery. Alkaline   
					batteries should have a minimum   
					capacity of 2500 mAh and NiMH   
					AA rechargeable batteries a   
					minimum capacity of 1500 mAh.   
					Use only a battery charger   
					officially recommended by the   
					battery manufacturer.   
					Pump casing   
					Shock, scratch, pharmaceutical   
					resistant plastic. All edges are   
					rounded.   
					*Determined using measurement methods according to IEC 60601-2-24-1998.   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					167   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Safety system   
					Alert system, beeps, information   
					on the display, vibrations, dual   
					microprocessor. The insulin pump is   
					controlled by two microprocessors.   
					The safety concept is based on one   
					processor (supervisor processor)   
					supervising the other (main   
					processor). Whenever a defect or   
					fault occurs in the main processor,   
					as identified by the supervisor   
					processor, the motor is immediately   
					switched off and an error E7:   
					ELECTRONIC ERROR occurs.   
					Conversely, the main processor   
					can also determine at any time   
					whether the supervisor is working   
					correctly. The brushless motor also   
					constitutes an important safety   
					component. It provides the best   
					possible reliability and accuracy in   
					insulin delivery.   
					Temperature ranges   
					Temporary Basal Rate   
					• During operation: 41°F to 104°F   
					(5°C to 40° C)   
					• Stored in its shipping case:   
					41°F to 113°F (5°C to 45° C)   
					Adjustable in 10% increments,   
					0-90% for decreases, 110-200%   
					for increases. The duration is   
					adjustable in 15-minute intervals,   
					up to a maximum of 24 hours. The   
					last programmed duration is given   
					by default for the next temporary   
					basal rate change programmed.   
					Weight   
					• Empty: approx. 2.8 oz. (80 g)   
					• Pump w/battery, full plastic   
					cartridge, and infusion set:   
					approx. 4 oz. (110 g)   
					Appendixes   
					168   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Technical Standards on Electromagnetic Emissions   
					The standard concerning electromagnetic compatibility of medical equipment   
					(IEC 60601-1-2) requires the concrete specification of the corresponding levels   
					which refer to specified electromagnetic interferences. The following is   
					Guidance and manufacturer’s declaration on electromagnetic emissions.   
					Emissions Test Compliance Electromagnetic Environment – Guidance   
					RF emissions   
					CISPR11   
					Group 1   
					The ACCU-CHEK Spirit insulin pump uses RF   
					energy only for its internal function.   
					Therefore, its RF emissions are very low   
					and are not likely to cause any interference   
					in nearby electronic equipment.   
					RF emissions   
					CISPR11   
					Class B   
					The ACCU-CHEK Spirit insulin pump is   
					suitable for use in establishments, including   
					domestic establishments and those directly   
					connected to the public low-voltage power   
					supply network that supplies buildings used   
					for domestic purposes.   
					Harmonic   
					emissions   
					IEC 61000-3-2   
					Not applicable   
					Not applicable   
					Voltage   
					fluctuations/   
					flicker emissions   
					IEC 61000-3-3   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					169   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Note 1   
					Note 2   
					At 80 and 800 MHz, the higher frequency range applies.   
					These guidelines may not apply in all situations. Electromagnetic   
					propagation is affected by absorption and reflection from structures,   
					objects and people.   
					a 
					Field strengths from fixed transmitters, such as base stations for radio   
					(cellular/cordless) telephones and land mobile radios, amateur radio,   
					AM and FM radio broadcast and TV broadcast, cannot be predicted   
					theoretically with accuracy. To assess the electromagnetic environment   
					due to fixed RF transmitters, an electromagnetic site survey should be   
					considered. If the measured field strength in the location in which the   
					ACCU-CHEK Spirit insulin pump is used exceeds the applicable RF   
					compliance level above, the ACCU-CHEK Spirit insulin pump should be   
					observed to verify normal operation. If abnormal performance is   
					observed, additional measures may be necessary, such as reorienting   
					or relocating the ACCU-CHEK Spirit insulin pump.   
					b 
					Over the frequency range 150 kHz to 80 MHz, field strengths   
					should be less than 10 V/m.   
					V1 in V:   
					Rated maximum 150 kHz to 80 MHz   
					output power out of ISM (Industrial,   
					of transmitter W Scientific, Medical Band)   
					Separation distance according to frequency of transmitter m   
					10.00   
					E1 in V/m:   
					80 MHz to   
					800 MHz   
					10.00   
					80 MHz to   
					2.5 GHz   
					0.01   
					0.1   
					1 
					10   
					100   
					0.04   
					0.11   
					0.35   
					1.11   
					3.50   
					0.04   
					0.11   
					0.35   
					1.11   
					3.50   
					0.07   
					0.22   
					0.70   
					2.21   
					7.00   
					Appendixes   
					170   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Technical Standards on Electromagnetic Immunity   
					Guidance and manufacturer’s declaration – electromagnetic immunity   
					The ACCU-CHEK Spirit insulin pump is intended for use in the electromagnetic   
					environment specified below. The customer or the user of the ACCU-CHEK   
					Spirit insulin pump should assure that it is used in such an environment.   
					IEC 60601   
					Immunity Test Test Level   
					Compliance   
					Level   
					Electromagnetic   
					Environment – Guidance   
					Electrostatic   
					discharge (ESD)   
					IEC 61000-4-2   
					6 kV contact 8 kV contact   
					8 kV air 15 kV air   
					Avoid any contact with   
					synthetic materials. A   
					(IEC 60601-2-24) relative humidity of 10% will   
					allow ESD to be more likely.   
					Power frequency 3 A/m 400 A/m 400 A/m   
					(50/60Hz)   
					magnetic field   
					Power frequency magnetic   
					fields should be at levels   
					characteristic of a typical   
					IEC 61000-4-8   
					(IEC 60601-2-24) location in a typical   
					commercial or hospital   
					environment.   
					Conducted RF 3 Vrms   
					IEC 61000-4-8 150 kHz to   
					80 MHz   
					Not applicable   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					171   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				IEC 60601 Compliance   
					Immunity Test Test Level Level   
					Radiated RF 3 V/m 10 V/m   
					Electromagnetic   
					Environment – Guidance   
					Portable and mobile RF   
					communications equipment   
					should be used no closer to   
					any part of the ACCU-CHEK   
					Spirit insulin pump, includ-   
					IEC 61000-4-3 80 MHz to 80 MHz   
					2.5 GHz 2.5 GHz   
					(IEC 60601-2-24) ing cables, than the recom-   
					mended separation distance   
					calculated from the equation   
					applicable to the frequency   
					of the transmitter.   
					Recommended separation   
					distance:   
					Ҁ 
					. 
					d = 0.4 ǰP 80 MHz   
					to 800 MHz   
					Ҁ 
					. 
					d = 0.7 ǰP 800 MHz   
					to 2.5 GHz   
					where P is the maximum   
					output power rating of the   
					transmitter in watts (W)   
					according to the transmitter   
					manufacturer and d is the   
					recommended separation   
					distance in meters (m).   
					Field strengths from fixed RF   
					transmitters, as determined   
					by an electromagnetic site   
					survey (a), should be less   
					than the compliance level in   
					each frequency range (b).   
					Interface may occur in the   
					vicinity of equipment   
					marked with the following   
					symbol:   
					Appendixes   
					172   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Accuracy of Flow Rate for Plastic Cartridges   
					Trumpet curve plotted from data after the end of the stabilization period   
					The trumpet curve shows the accuracy of the delivery rate in relation to the   
					observation period.   
					The measurements were made at a medium basal rate of 1.0 U/h according to   
					IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room   
					temperature.   
					The average deviation of the delivered amount (overall mean percentage flow   
					error) for U100 insulin is ≤ 5%.   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					173   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Start-up graph over the stabilization period   
					The start-up graph shows changes in the flow rate over the stabilization time.   
					The measurements were made at a medium basal rate of 1.0 U/h according to   
					IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room   
					temperature.   
					Influence of height on delivery accuracy   
					The average deviation of the delivered amount (overall mean percentage flow   
					error) for U100 insulin is ≤ 10% when your ACCU-CHEK Spirit insulin pump is   
					located 3.3 feet (1 m) above/below the level of infusion site.   
					Appendixes   
					174   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Trumpet curve plotted from data after the end of the stabilization period   
					The trumpet curve shows the accuracy of the delivery rate in relation to the   
					observation period.   
					The measurements were made at a medium basal rate of 0.1 U/h according to   
					IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room   
					temperature.   
					The average deviation of the delivered amount (overall mean percentage flow   
					error) for U100 insulin is ≤ 20%.   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					175   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Start-up graph over the stabilization period   
					The start-up graph shows changes in the flow rate over the stabilization time.   
					The measurements were made at a medium basal rate of 0.1 U/h according to   
					IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room   
					temperature.   
					Bolus   
					For U100 insulin the delivery accuracy of a maximum bolus is ≤ 5% and the   
					delivery accuracy of a minimum bolus is ≤ 30%.   
					The measurements were made according to IEC 60601-2-24:1998 with a   
					Disetronic Classic infusion set PC 16/110 at room temperature.   
					Delivery Mode   
					Delivery Rate/Volume   
					Accuracy   
					Range   
					Basal   
					Bolus   
					0.1 U/h   
					0.1 U   
					20%   
					30%   
					0.08 - 0.12 U/h   
					0.07 - 0.13 U   
					Appendixes   
					176   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Appendix B: Configuration Parameters   
					Your pump is built by Disetronic Medical Systems with a standard configuration   
					(factory settings). A list containing the important factory settings is added to the   
					original insulin pump package when your pump is first shipped. You may need   
					to adapt the factory settings to meet your individual needs. Set the configuration   
					parameters directly on your insulin pump. Always consult your doctor or   
					healthcare team before changing parameters.   
					The list on the next two pages gives you the full range for the adjustable   
					configuration parameters on your pump. Also included are examples of typical   
					range and parameter settings.   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					177   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Typical Range and   
					Parameter Settings   
					Maximum Parameter   
					Range   
					Basal Rate   
					– hourly basal rate +   
					max. TBR   
					U100: 62.5 IU/h Insulin Type Max. Amount   
					U100 0 – 62.50 IU/h   
					(hourly basal rate   
					combined with temporary   
					basal rate, max. increase)   
					– hourly basal rates   
					increment   
					0.1 U   
					– hourly basal rates   
					U100: 10.0 IU/h Insulin Type Max. Amount   
					U100 0 – 25.0 IU/h   
					– temporary basal rate   
					– TBR increase or decrease   
					in increments of   
					– TBR duration   
					– TBR duration increments   
					100%   
					0 – 200%   
					10%   
					max. 24 h   
					15 min   
					10%   
					15 min – 24 h   
					15 min, 30 min, 1 h   
					– basal rate profiles available   
					– basal rate lock   
					5 
					off   
					1 – 5   
					on or off   
					Bolus   
					– bolus amount per   
					bolus delivery   
					U100: 25.0 IU/h Insulin Type Max. Amount   
					U100 0 – 25.0 U   
					– bolus amount increments   
					for “Quick” and “Scroll”   
					Standard Bolus   
					0.5U   
					0.1, 0.2, 0.5, 1.0 or 2.0 U   
					– bolus duration (Extended   
					bolus, MultiWave bolus)   
					– bolus duration increments   
					max. 24 h   
					15 min   
					15 min – 24 h   
					15 min, 30 min, 1h   
					Appendixes   
					178   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Typical Range and   
					Parameter Settings   
					Maximum Parameter   
					Range   
					Prime volume   
					U100: 25.0 U   
					Insulin Type   
					U100   
					Max. Amount   
					0 – 25.0 IU/h   
					Display orientation   
					Display contrast   
					standard   
					medium   
					standard or inverted   
					13 steps   
					User menus   
					STANDARD   
					3 (STANDARD, ADVANCED   
					or CUSTOM)   
					Volume (for beeps)   
					maximum   
					off, 1 – 4 steps   
					Alarm signals turned on beeps and vibrations   
					beeps, vibrations or both   
					Battery type   
					AA battery   
					AA Alkaline or   
					AA NiMH rechargeable battery   
					KeyLock   
					off   
					off   
					turned on or off   
					Automatic off   
					turned on or off   
					0.1 – 24h   
					Alarm clock   
					off   
					turned on or off   
					when on: once or every   
					day’s occurrence   
					Time format   
					Date format   
					American (12 h am/pm) or   
					European (24 h)   
					American (mm/dd/yy) or   
					European (dd.mm.yy)   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					179   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Appendix C: Abbreviations   
					approx.   
					BR   
					Approximately   
					Basal rate   
					BRP   
					DDM   
					h 
					Basal rate profile   
					Diabetes Data Management   
					Hour(s)   
					IEC   
					International Electrotechnical Commission   
					incl.   
					IR   
					Inclusive   
					Infrared   
					I.U., IU   
					International Units in context with biological effectiveness of   
					a certain insulin amount   
					kPa   
					LCD   
					LED   
					Kilo pascals   
					Liquid Crystal Display   
					Light Emitting Diode for infrared communication with, for   
					example, a personal computer (PC)   
					NiMH   
					rel.   
					Nickel Metal Hydride (used with rechargeable batteries)   
					Relative   
					sec.   
					Second(s)   
					TBR   
					Temporary Basal Rate   
					U 
					see I.U.   
					U/h   
					Amount of International Units of insulin delivered per hour   
					Each milliliter of liquid contains 100 International Units of insulin   
					American date format: month/day/year   
					European date format: day.month.year   
					U100   
					mm/dd/yy   
					dd.mm.yy   
					Appendixes   
					180   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Appendix D: Beeps and Melodies   
					You can turn off either the pump beeps or the vibrations, but not both   
					simultaneously. Please refer to sections “Adjust the Beep Volume”(page 116),   
					and “Setting Up Alarm Signals” (page 117) for more information.   
					Each press of   
					Each press of   
					Each press of   
					confirms each programmed bolus   
					increment when programming a “Quick” Standard Bolus   
					Each press of confirms the STOP-Warning   
					Temporary Basal Rate is active   
					Unlock the KeyLock in RUN mode, enter the RUN or INFO   
					screen, start a Temporary Basal Rate   
					Lock the KeyLock in RUN mode   
					Unlock the KeyLock in STOP mode, enter STOP screen   
					Lock the KeyLock in STOP mode   
					Exit a screen   
					Cancel a running bolus or stop priming   
					Start-up procedure successful   
					Copy an hourly basal rate   
					Alert and error   
					Maximum amount reached   
					Minimum amount reached   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					181   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Appendix E: Vibrations   
					Your pump signals alerts and errors using beeps and vibrations. You can turn off   
					either the beeps or the vibrations, but you cannot turn off both simultaneously.   
					When you program a “Quick” Standard Bolus, your ACCU-CHEK Spirit insulin pump   
					confirms all programming by vibrating. You cannot turn off these vibrations.   
					Your ACCU-CHEK Spirit insulin pump vibrates when   
					l (Short vibration)   
					Your ACCU-CHEK Spirit insulin pump performs a   
					start-up procedure   
					(Medium vibration)   
					Your ACCU-CHEK Spirit insulin pump confirms   
					the programmed bolus amount of the “Quick”   
					Standard bolus   
					(Long vibration)   
					You reset the “Quick” Standard Bolus amount to   
					zero, you cancel the “Quick” Standard Bolus, or   
					when the STOP-Warning occurs   
					See the sections “Programming a Standard Bolus” (page 77) and “Setting Up   
					Alarm Signals” (page 117) for more information.   
					Appendixes   
					182   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Appendix F: Symbols   
					General Symbols   
					Read user guide   
					Sterilized using ethylene oxide   
					Sterilized using radiation   
					Sterilized using antiseptics   
					Year of manufacture   
					Batch number   
					Expiration date   
					Item number   
					Serial number   
					Admissible temperature range during operation   
					Admissible temperature range   
					Admissible humidity range   
					Admissible air pressure range   
					Fragile – handle with care   
					Protect against moisture   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					183   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Protect from heat and sunlight   
					For single use only   
					Does not contain pyrogen   
					Does not contain PVC   
					Recycling   
					Do not throw away   
					Do not use if the package is damaged   
					Flammable   
					See   
					Electronic device of type BF according to the standard   
					IEC 60601-1. Protection against electrical shock.   
					Symbol for protection against the effect of temporary immersion   
					in water [up to 60 minutes and 2.5 meters (8 feet)], according   
					to IEC 60529.   
					Marking of conformity according to the European Medical   
					Device Directive MDD 93/42/EEC with the number of the   
					notified body.   
					Federal law (USA) restricts this device to sale by or on the   
					order of a physician.   
					Federal Communications Commission, part 15 on radio   
					frequency devices.   
					Manufacturer   
					Appendixes   
					184   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Symbol Definition   
					Start your ACCU-CHEK Spirit insulin pump   
					STOP screen and Stop insulin delivery   
					Time, and Set the time and date menu   
					Date   
					Locked keys when the KeyLock function is turned on   
					Unlocked keys when the KeyLock function is turned on   
					Change the cartridge menu and cartridge content   
					Cartridge low warning   
					Empty cartridge   
					Low battery   
					Empty battery   
					Alkaline AA battery   
					Rechargeable AA battery   
					Amount of International Units of insulin delivered per hour   
					Basal rate profile   
					Temporary Basal Rate   
					TBR percentage   
					Daily basal rate total   
					Basal rate programming unlocked   
					Basal rate programming locked   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					185   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Standard Bolus and Bolus increment programming   
					Extended Bolus and Extended delivery of the MultiWave Bolus   
					MultiWave Bolus   
					Immediate bolus delivery of MultiWave Bolus   
					Decreased Temporary Basal Rate (0–90%)   
					Increased Temporary Basal Rate (110–200%)   
					Remaining duration of an ongoing Extended Bolus, MultiWave   
					Bolus, TBR, or automatic off   
					Daily total of insulin delivered (basal rate plus boluses)   
					Prime the infusion set (menu)   
					Alert or error occurring   
					Alert   
					Error   
					Information menu   
					Pump timer   
					Select the user menus STANDARD, ADVANCED or CUSTOM   
					SETUP MENU STANDARD menu   
					SETUP MENU ADVANCED menu   
					Beeps turned on   
					Vibrations turned on   
					Beeps and vibrations turned on   
					Appendixes   
					186   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Set beep volume   
					Screen orientation   
					European date format   
					American date format   
					Alarm clock turned on   
					Alarm clock turned off   
					Single alarm when setting an alarm clock   
					Every day repeat alarm when setting an alarm clock   
					PC communication   
					Set beep volume   
					Check key   
					Up key   
					Down key   
					Timeout of a menu   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					187   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Appendix G: Sterile Products and Accessories   
					Sterile Products   
					Cartridges   
					Name   
					Comments   
					Empty ACCU-CHEK   
					3.15ml Plastic Cartridge   
					Plastic cartridges are single-use   
					items. Do not reuse cartridges.   
					Refer to the instructions of the   
					insulin you are using for information   
					of proper use and acceptable   
					temperature range for storage   
					and handling.   
					ACCU-CHEK infusion sets   
					Name   
					Comments   
					ACCU-CHEK Ultraflex   
					The ACCU-CHEK Ultraflex infusion   
					set has a specially tapered cannula   
					and reversible connector for easy,   
					comfortable insertion and wearing   
					convenience.   
					ACCU-CHEK Tender   
					ACCU-CHEK Rapid-D   
					Variable insertion angle and choice   
					of cannula for maximum control   
					and comfort.   
					Ultra-fine stainless needle inserts   
					easily with one hand and is a great   
					option for new pumpers or those   
					with vision or dexterity issues.   
					Appendixes   
					188   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Accessories   
					Name   
					Comments   
					Adapter   
					Replace your adapter with every   
					10th cartridge.   
					Battery   
					If used in a typical usage pattern   
					[50 U/day using U100 insulin;   
					operating temp. of 72°F 6°F   
					(22°C 3°C)] the battery life is   
					approx. four weeks for alkaline.   
					One 1.5 AA alkaline battery with a   
					minimum capacity of 2500mAh   
					Rechargeable Battery   
					One AA NiMH rechargeable battery with   
					a minimum capacity of 1500 mAh   
					If used in a typical usage pattern   
					[50 U/day using U100 insulin;   
					operating temp. of 72°F 6°F   
					(22°C 3°C)] the battery life is   
					approx. one week for rechargeable   
					batteries.   
					Battery Cover   
					Replace your battery cover with   
					every 4th battery.   
					Carrying Systems   
					Disetronic Medical Systems offers   
					a wide range of carrying system   
					solutions to best fit your lifestyle.   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					189   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Appendix H: Glossary   
					Adapter   
					The adapter connects the cartridge and infusion set. It has two seals for the   
					cartridge compartment of your insulin pump. The two small vents on the   
					adapter let air pressure equalize.   
					Basal rate   
					The amount of insulin needed to cover your own basic insulin needs. In insulin   
					pump therapy, your basal rate is determined by your doctor or healthcare team   
					and can be adjusted to meet your body’s specific needs throughout the day.   
					Your basal rate is delivered by your insulin pump according to your personal   
					basal rate profile or profiles.   
					Basal rate profile   
					Your pump can be set to deliver up to five different basal rate profiles to match   
					changing insulin needs (for example, during the week and on the weekend). A   
					basal rate profile consists of 24 programmed hourly basal rates.   
					Basal rate total   
					The sum of all 24 hourly basal rates in one basal rate profile is called the (daily)   
					basal rate total.   
					Bolus   
					The amount of insulin delivered (in addition to the basal rate) to cover the   
					intake of food and correct high blood glucose levels. The bolus amount is   
					determined by your doctor or healthcare team’s guidelines, your blood glucose   
					level, your food intake, and your activity level.   
					Carrying system   
					A variety of special carrying systems made of different materials are designed   
					to suit your individual needs while wearing your pump.   
					Cartridge   
					The insulin reservoir of your pump. It holds 3.15 ml of rapid-acting insulin analog.   
					When using U100 insulin, 3.15 ml is equal to 315 International Units of insulin.   
					Cartridge compartment   
					The opening in your pump for the cartridge.   
					Appendixes   
					190   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Daily insulin total   
					The total amount of insulin delivered (basal rate plus boluses) in a 24-hour day,   
					beginning at midnight. This amount does not include any insulin needed for the   
					priming of infusion sets.   
					Diabetes Data Management (DDM)   
					Diabetes Data Management is the recording of the therapy-relevant data   
					gained from your insulin delivery system and your blood glucose measurement   
					system (such as an ACCU-CHEK blood glucose monitor) in order to analyze and   
					illustrate this data on a PC or other communication device.   
					Factory settings   
					The standard configuration (“factory settings”) on a new pump. These settings   
					can be adapted to your individual needs directly on your pump.   
					Filling aid   
					An accessory that can be used to help you fill an empty ACCU-CHEK 3.15 ml   
					plastic cartridge from your vial of insulin.   
					Free flow   
					Free flow of insulin from the infusion set may occur when two conditions are   
					met: first, if the cartridge with a connected infusion set and the piston rod of   
					your pump are not correctly coupled and second, if you position your pump at   
					a higher level than the infusion site.   
					Headset (also known as cannula housing)   
					The part of an infusion set that includes the soft cannula or needle. The   
					infusion set tubing connects either directly to the headset, or connects to a   
					shorter length of tubing permanently attached to the headset.   
					Hourly basal rate   
					The amount of insulin delivered by your pump in 3-minute intervals for a   
					given hour.   
					Infusion set   
					Connects your pump to your body. The insulin is delivered from the cartridge   
					through the infusion set tubing and cannula or needle into your subcutaneous   
					tissue.   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					191   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Infusion site   
					The place where the infusion set cannula or needle is inserted into your   
					subcutaneous tissue to deliver insulin.   
					Insulin   
					A hormone that helps cells turn glucose into energy. Insulin is produced in the   
					beta cells of the pancreas (also called islets of Langerhans).   
					Insulin, rapid-acting (insulin analog)   
					A type of insulin created using recombinant DNA technology. The onset time   
					of insulin analog is 5 to 15 minutes.   
					IPX8 according to IEC 60529   
					Protection against the effects of temporary immersion in water. Ingress of water   
					in quantities causing harmful effects is not possible when the device is temporarily   
					immersed in water under standardized conditions [maximum of 60 minutes a   
					day and 8 feet (2.5 meters) depth].   
					Luer-lock connector   
					A standardized fitting at the end of the infusion set, the cartridge, and the adapter   
					that lets them be locked together without leaking.   
					Personal settings   
					Your pump must be programmed with your personal settings prior to starting   
					insulin pump therapy. Personal settings include the basal rate profile(s), the   
					correct time and date, and all other variable values that can be tailored to   
					your individual needs on your pump.   
					RUN mode   
					During normal use, your pump is in RUN mode and insulin is continuously   
					delivered. Boluses, Temporary Basal Rates, and almost all functions can be   
					programmed from RUN mode.   
					RUN screen   
					The RUN screen is the starting point for all functions that can be accessed   
					while your pump delivers insulin. Your pump displays the RUN screen during   
					normal use when no programming is currently being performed. The time,   
					current hourly basal rate, basal rate profile, and all functions currently   
					activated appear on the RUN screen.   
					Appendixes   
					192   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Settings   
					Settings are individually programmable values and parameters that affect the   
					way your pump works.   
					Scrolling   
					Scrolling allows fast and easy programming of larger values.   
					STOP mode   
					When your pump is in STOP mode, it does not deliver any insulin. The insulin   
					delivery is only stopped if an error occurs or if your pump is put into STOP   
					mode (for example, when changing the cartridge, adapter or infusion set, or   
					for data transfer). Functions such as Extended Bolus or Temporary Basal Rate   
					are interrupted by putting your pump into STOP mode.   
					STOP screen   
					The STOP screen is the starting point for all functions that require that your   
					pump does not deliver insulin.   
					Temporary Basal Rate   
					Temporary increase or decrease of your basal rate profile in percentages (from   
					0 – 200%) to match changing insulin needs due to increased or decreased   
					activity level, illness, or stress.   
					Timeout   
					For your safety and convenience, your pump returns to the RUN or STOP   
					screen if no key is pressed within a certain time. Any changes made are not   
					saved.   
					U100   
					Each milliliter of liquid contains 100 International Units of insulin. Your pump   
					was developed only to use U100 rapid-acting insulin analogs.   
					Appendixes   
					ACCU-CHEK Spirit Insulin Pump   
					193   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				LIMITED FOUR-YEAR WARRANTY   
					DISETRONIC MEDICAL SYSTEMS, INC. (“DISETRONIC”) WARRANTS THE ACCU-CHEK SPIRIT   
					INSULIN PUMP (“INSULIN PUMP”) AGAINST DEFECTS IN MATERIALS AND WORKMANSHIP FOR A   
					PERIOD OF FOUR (4) YEARS FROM THE DATE OF PURCHASE. PROOF OF PURCHASE DATE MAY   
					BE REQUIRED. THIS WARRANTY DOES NOT INCLUDE SUPPLIES AND ACCESSORIES, INCLUDING   
					BUT NOT LIMITED TO, CARTRIDGES, BATTERIES, OR INFUSION SETS.   
					DURING THE WARRANTY PERIOD, DISETRONIC WILL REPAIR OR REPLACE, IN ITS SOLE   
					DISCRETION, ANY INSULIN PUMP THAT DOES NOT WORK PROPERLY BECAUSE OF A DEFECT IN   
					MATERIALS OR WORKMANSHIP. THE PURCHASER’S SOLE AND EXCLUSIVE REMEDY WITH   
					RESPECT TO THE INSULIN PUMP SHALL BE REPAIR OR REPLACEMENT. THIS WARRANTY APPLIES   
					ONLY TO NEW DEVICES. IN THE EVENT A PUMP IS REPAIRED OR REPLACED, THE WARRANTY   
					PERIOD WILL NOT BE EXTENDED.   
					THIS WARRANTY IS VALID ONLY IF THE INSULIN PUMP IS USED IN ACCORDANCE WITH THE USER   
					INSTRUCTIONS. THIS WARRANTY WILL NOT APPLY:   
					1. IF DAMAGE RESULTS FROM CHANGES OR MODIFICATIONS MADE TO THE INSULIN PUMP BY   
					THE USER OR THIRD PERSONS AFTER THE DATE OF MANUFACTURE;   
					2. IF DAMAGE RESULTS FROM SERVICE OR REPAIRS PERFORMED BY ANY PERSON OR ENTITY   
					OTHER THAN DISETRONIC.   
					3. IF DAMAGE RESULTS FROM A FORCE MAJEURE OR OTHER EVENT BEYOND THE CONTROL OF   
					DISETRONIC; OR   
					4. IF DAMAGE RESULTS FROM NEGLIGENCE OR IMPROPER USE, INCLUDING BUT NOT LIMITED   
					TO IMPROPER STORAGE, PHYSICAL ABUSE SUCH AS DROPPING, OR OTHERWISE.   
					THIS WARRANTY SHALL BE PERSONAL TO THE ORIGINAL PURCHASER. ANY SALE, RENTAL, OR   
					OTHER TRANSFER OR USE OF THE INSULIN PUMP COVERED BY THIS WARRANTY TO OR BY A   
					USER OTHER THAN THE ORIGINAL PURCHASER SHALL CAUSE THIS WARRANTY TO IMMEDIATELY   
					TERMINATE.   
					THIS WARRANTY IS EXCLUSIVE OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED INCLUDING   
					WITHOUT LIMITATION THE IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A   
					PARTICULAR PURPOSE. IN NO EVENT SHALL DISETRONIC OR ITS SUPPLIERS OR DISTRIBUTORS   
					BE LIABLE FOR ANY INCIDENTAL, CONSEQUENTIAL, INDIRECT, SPECIAL OR PUNITIVE DAMAGES   
					ARISING FROM OR IN ANY WAY CONNECTED WITH THE PURCHASE OR USE OF THE INSULIN   
					PUMP.   
					Appendixes   
					194   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Cartridge; 9, 18   
					Changing; 73-74   
					Index   
					Connecting; 47-48   
					Filling; 43-46   
					Inserting; 49-52   
					Low; 149, 185   
					Accessories; 17-23, 189   
					Adapter; 9, 20, 47-48, 190   
					Advanced User Menu; 99, 122-128   
					Air Bubbles; 45, 53, 145   
					Alarm Clock; 127-128, 151   
					Alarm History; 109   
					CHECK Key; 9, 187   
					Cleaning; 141   
					Copying Hourly Basal Rate; 40-41   
					Alarm Signals; 16, 117   
					Alcohol Consumption; 144   
					Alerts and Errors; 147-164, 199-200   
					Automatic Off; 118, 156   
					Data Interrupted; 163   
					Data Transfer; 104   
					Date; 32-34, 150   
					Date Format; 130   
					Daily Insulin Totals; 110, 191   
					Display; 9-10   
					Contrast; 135   
					Orientation; 120, 187   
					Disposal; 140   
					Backlight; 10   
					Basal Rate Profile; 36, 123, 126, 190   
					Lock; 133   
					Programming; 37-41, 124   
					Selecting; 125   
					Temporary; 93-96   
					Battery; 21, 141   
					DOWN Key; 9, 187   
					Cover; 22   
					Depleted; 155   
					Inserting or changing; 25-27   
					Key; 23, 71   
					Life; 165, 189   
					Electromagnetic; 61, 169-172   
					Electronic Error; 160   
					Emergency Kit; 23   
					Errors; 147-148, 154-164   
					Exit Options; 102   
					Low; 150, 185   
					Type; 119, 167   
					Beeps; 15, 116, 181, 186-187   
					Bolus; 75   
					Filling Aid; 43   
					Guarantee; 4   
					Extended; 83-87   
					History; 107-108   
					MultiWave; 88-92   
					“Quick” Standard; 77-80   
					“Scroll” Standard; 80-82   
					History; 107   
					Alarm; 109   
					Bolus; 107-108   
					Daily Insulin; 110   
					Temporary Basal Rate; 111   
					Index   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
					195   
				Immunity; 171-172   
					Infusion Set; 9, 19-20   
					Power Interrupt; 161   
					Prime Quantity; 132   
					Priming; 53-55   
					Changing the; 70-72   
					Not Primed; 162-163   
					Priming; 53-55   
					Infusion Site; 56-57, 72   
					Insulin Delivery;   
					Starting; 65   
					Products, Sterile; 17-18, 188   
					Profile, See Basal Rate Profile   
					Programming;   
					Additional Basal Rate   
					Profile; 124   
					Stopping; 66   
					Insulin Pump Therapy;   
					Basal Rate Profile; 37   
					Extended Bolus; 83-87   
					MultiWave Bolus; 88-92   
					“Quick” Standard Bolus;   
					77-80   
					Interruption of; 68-69   
					Resuming; 69   
					KeyLock; 13   
					“Scroll” Standard Bolus;   
					80-82   
					Temporary Basal Rate; 93-96   
					Pump; 9   
					Turning On; 113-114   
					Turning Off; 113-114   
					Unlocking; 181, 185   
					Keys; 11   
					Stopping; 66   
					Key Combinations; 12   
					Storing; 140   
					Wearing; 59-60   
					Language; 134   
					Language Error; 164   
					Liquids; 64   
					Quick Info Screen; 14   
					Quick Information; 30   
					Luer-Lock; 192   
					Repair; 142   
					Maintenance; 141   
					Mechanical Error; 159-160   
					Menus; 99-102   
					RUN Screen; 13, 192   
					Safety System; 147, 168   
					Screen; 9-10   
					Advanced; 122   
					Custom; 136   
					Scrolling; 193   
					Standard; 103   
					Setup Menu;   
					MultiWave Bolus;   
					Programming; 88-91   
					Cancelling; 91-92   
					Advanced; 129, 186   
					Standard; 113, 186   
					Standard User Menu; 99, 103   
					Start-Up; 28-31   
					Occlusion; 146   
					Error; 156-158, 167   
					Orientation; 120   
					Status Screen; 15   
					Sterile; 17, 188   
					Stop-Warning; 16, 67   
					Index   
					196   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Stopping;   
					Insulin Delivery; 66   
					Quick Reference Table for Errors   
					Priming; 55   
					System Check; 28-31   
					No. Error   
					What You Need to Do   
					E1   
					E2   
					E3   
					CARTRIDGE EMPTY   
					Change the cartridge.   
					Change the battery.   
					Temporary Basal Rate;   
					Cancelling; 97   
					History; 111   
					BATTERY DEPLETED   
					AUTOMATIC OFF   
					If necessary, put your ACCU-CHEK Spirit in   
					RUN mode.   
					Over; 96, 153   
					Programming; 93-96   
					Time and Date; 32   
					American; 32, 130   
					European; 32, 130   
					Time and Date, Review; 150   
					E4   
					E5   
					OCCLUSION   
					Immediately check your blood glucose level,   
					because insulin delivery has been interrupted.   
					If your blood glucose level is high, take   
					appropriate actions according to your doctor’s   
					or healthcare team’s instructions. See the   
					section “Error E4: OCCLUSION.”   
					Time Remaining; 112   
					Troubleshooting; 143-146   
					Tubing; 20, 132   
					END OF OPERATION   
					Arrange for immediate replacement of your   
					insulin pump. Contact Pump   
					Support at   
					1-800-688-4578 to dis-   
					cuss your options for further use of the   
					ACCU-CHEK Spirit insulin pump. Contact   
					your healthcare professional for alternative   
					insulin therapy options.   
					UP Key; 9, 187   
					User Menu; 99-102   
					Advanced; 122   
					Custom; 136   
					E6   
					MECHANICAL ERROR   
					ELECTRONIC ERROR   
					POWER INTERRUPT   
					CARTRIDGE ERROR   
					See the section “Error E6: MECHANICAL   
					ERROR.”   
					Standard; 103   
					E7   
					See the section “Error E7: ELECTRONIC   
					ERROR.”   
					Water; 63   
					Contact with; 63   
					After Contact with; 64   
					E8   
					See the section “Error E8: POWER INTER-   
					RUPT.”   
					E10   
					See the section “Error E10: CARTRIDGE   
					ERROR.”   
					E11   
					E12   
					E13   
					SET NOT PRIMED   
					DATA INTERRUPTED   
					LANGUAGE ERROR   
					Prime your infusion set.   
					Restart the data transfer.   
					See the section “Error E13: LANGUAGE   
					ERROR.”   
					Index   
					ACCU-CHEK Spirit Insulin Pump   
					197   
					Quick Reference Table for Errors   
					200   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Quick Reference Table for Alerts   
					Diagram of User Menus   
					In all cases, turn off and confirm the alert prior to addressing the issue.   
					Your Insulin pump STANDARD MENU   
					No. Alert   
					What You Need to Do   
					KeyLock   
					A1   
					CARTRIDGE LOW   
					Change cartridge before it is completely   
					empty.   
					quick info   
					Standard   
					bolus   
					KeyLock   
					quick info   
					«Scroll» Stan-   
					dard bolus   
					temporary   
					basal rate   
					stop your   
					information   
					menu   
					program   
					set the time   
					and date   
					setup menu   
					standard   
					select the   
					user menu   
					A2   
					A3   
					A4   
					BATTERY LOW   
					Replace the battery as soon as possible.   
					change the   
					cartridge   
					prime the   
					communi-   
					cation   
					Accu-Chek   
					profile no.1   
					infusion set   
					Standard   
					bolus   
					REVIEW TIME AND DATE Set the time and date.   
					bolus   
					amount   
					percentage   
					duration   
					bolus   
					history   
					daily basal   
					rate total   
					set the   
					hour   
					KeyLock   
					user menu   
					start your   
					remove the   
					cartridge   
					start   
					priming   
					data transfer   
					to PC   
					Accu-Chek   
					ALARM CLOCK   
					Put your ACCU-CHEK Spirit in RUN mode,   
					if necessary.   
					alarm   
					history   
					1st hour   
					2nd hour   
					… 
					set the   
					minute   
					beep   
					volume   
					BR profile   
					no.…   
					returning the   
					piston rod   
					stop   
					priming   
					(A) «Quick»   
					with direct access from the RUN screen   
					Standard bolus using the a and s keys of your   
					insulin pump.   
					daily insulin   
					totals   
					set the   
					year   
					alarm   
					signals   
					fill volume   
					315 units   
					A5   
					A6   
					PUMP TIMER   
					Your pump’s operating time will expire   
					soon. Go to the INFORMATION screen on   
					your pump, then move to the PUMP TIMER   
					screen. The days until the pump timer   
					expires is shown. Make arrangements for   
					replacement of the pump before the timer   
					reaches zero days.   
					(B) «Scroll»   
					menu-guided using the STANDARD BOLUS   
					TBR history   
					set the   
					month   
					automatic   
					off   
					start-up   
					procedure   
					Standard bolus menu and the scroll function of the a and s   
					keys for programming the bolus amount.   
					maintenance   
					24th hour   
					set the   
					day   
					battery   
					type   
					For a complete review of your insulin pump features as well as   
					the warnings and precautions associated with its use, please   
					refer to your ACCU-CHEK Spirit Insulin Pump User Guide.   
					display   
					orientation   
					Your Insulin pump ADVANCED MENU   
					KeyLock   
					TBR CANCELLED   
					Put your ACCU-CHEK Spirit insulin pump in   
					RUN mode, if necessary. Ensure the can-   
					cellation was intended and program a new   
					Temporary Basal Rate, if necessary.   
					quick info   
					Standard   
					bolus   
					KeyLock   
					quick info   
					«Scroll» Stan-   
					dard bolus   
					Extended   
					bolus   
					MultiWave   
					bolus   
					temporary   
					basal rate   
					stop your   
					Accu-Chek   
					information   
					menu   
					change your   
					BR profile   
					program 
					profileno.12345   
					alarm clock   
					set the time   
					and date   
					setup menu   
					standard   
					setup menu   
					advanced   
					select the   
					user menu   
					change the   
					cartridge   
					prime the   
					infusion set   
					communi-   
					cation   
					bolus   
					amount   
					bolus   
					amount   
					bolus   
					amount   
					percentage   
					duration   
					bolus   
					history   
					BR profile   
					no. …   
					daily basal   
					rate total   
					alarm   
					occurance   
					set the   
					hour   
					KeyLock   
					time   
					format   
					user menu   
					start your   
					Accu-Chek   
					remove the   
					cartridge   
					start   
					priming   
					data transfer   
					to PC   
					A7   
					A8   
					TBR OVER   
					Decide whether a further Temporary Basal   
					Rate change is appropriate and program   
					one, if necessary.   
					bolus   
					duration   
					immediate   
					bolus   
					alarm   
					history   
					1st hour   
					2nd hour   
					… 
					alarm   
					hour   
					set the   
					minute   
					beep   
					volume   
					date   
					format   
					BR profile   
					no.…   
					returning the   
					piston rod   
					stop   
					priming   
					bolus   
					duration   
					fill volume   
					315 units   
					daily insulin   
					totals   
					alarm   
					minute   
					set the   
					year   
					alarm   
					signals   
					bolus   
					increment   
					BOLUS CANCELLED   
					Put your ACCU-CHEK Spirit insulin pump in   
					RUN mode, if necessary. Ensure that the   
					cancellation was intended and program a   
					new bolus, if necessary.   
					TBR history   
					set the   
					month   
					automatic   
					off   
					prime   
					quantity   
					start-up   
					procedure   
					maintenance   
					24th hour   
					set the   
					day   
					battery   
					type   
					basal   
					rate lock   
					display   
					orientation   
					language   
					display   
					contrast   
					Quick Reference Table for Alerts   
					ACCU-CHEK Spirit Insulin Pump   
					199   
					Index   
					198   
					ACCU-CHEK Spirit Insulin Pump   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 |