Gossen MetraWatt PROFITEST EMOBILITY Operating Instructions User Manual

Operating Instructions  
PROFITESTEMOBILITY  
Adapter for Standards-Compliant Testing of Single and 3-Phase,  
Mode 2 and 3 Charging Cables  
3-349-981-15  
7/2.24  
Safety Instructions  
1
Safety Instructions  
Read and follow these instructions carefully  
and completely in order to ensure safe and  
proper use.  
from operation permanently and secure  
it against inadvertent use.  
If there are any signs of interior damage  
to the instrument or accessories (e.g.  
loose parts in the housing), permanently  
remove the instrument/accessories from  
operation and secure them against inad-  
vertent use.  
Route cables in an orderly fashion, e.g.  
the mains power cable. Loose, disor-  
derly cables result in unnecessary dan-  
ger of tripping and falling.  
The instrument may only be used for the  
measurements/tests described in the  
documentation for the instrument.  
The instrument may only be connected to  
electrical systems with a maximum of 230/  
400 V which comply with applicable  
safety regulations (e.g. IEC 60346, VDE  
0100) and are protected with a fuse or cir-  
cuit breaker with a maximum rating of  
16 A.  
The test adapter may only be used for  
testing mode 2 and 3 charging cables.  
No power consuming devices may be  
connected to any of the sockets.  
Measurements within electrical systems  
are prohibited.  
Read the operating instructions for the  
respective test instrument as well, in particu-  
lar the sections concerning the R  
ISO  
PE, R  
and time-to-trip measurements, as well as  
the tripping test.  
General  
Measurements/tests may only be per-  
formed by a qualified electrician, or  
under the supervision and direction of a  
qualified electrician. The user must be  
instructed by a qualified electrician con-  
cerning performance and evaluation of  
the tests/measurements.  
Observe and comply with all safety regu-  
lations which are applicable for your  
work environment.  
Wear suitable and appropriate personal  
protective equipment (PPE) whenever  
working with the instrument.  
Handling  
The instruments and accessories of  
Gossen Metrawatt GmbH are designed  
such as to ensure optimum compatibility  
with the Gossen Metrawatt GmbH prod-  
ucts that are expressly provided for  
them. Unless otherwise expressly con-  
firmed in writing by Gossen Metrawatt  
GmbH, they are not intended and suited  
for use with other products.  
Observe the manufacturer’s specifica-  
tions for DUTs.  
Operating Conditions  
Do not use the instrument and its acces-  
sories after long periods of storage under  
unfavorable conditions (e.g. humidity,  
dust or extreme temperature).  
Do not use the instrument and its acces-  
sories after extraordinary stressing due  
to transport.  
The instrument must not be exposed to  
direct sunlight.  
Only use the instrument and its accesso-  
ries within the limits of the specified tech-  
nical data and conditions (ambient con-  
ditions, IP protection code, measuring  
category etc.).  
Use the instrument in undamaged condi-  
tion only.  
Inspect the instrument before use. Pay  
particular attention to damage, inter-  
rupted insulation or kinked cables.  
Also test the LEDs: if the LEDs are  
defective, voltage at the ICCB-IN1 sock-  
ets is no longer indicated.  
If the instrument doesn’t function flaw-  
lessly, permanently remove it from oper-  
ation and secure it against inadvertent  
use.  
If the instrument is damaged during use,  
for example if it’s dropped, remove it  
3 | 22  
 
Application  
Measurement Cables and Establishing Contact  
2
Application  
Plugging in the measurement cables  
must not necessitate any undue force.  
Never touch conductive ends (e.g. of  
test probes).  
Do not touch the insulation measuring  
instrument’s test probes during insula-  
tion resistance measurements!  
2.1  
Intended Use /  
Use for Intended Purpose  
The PROFITEST EMOBILITY is an adapter  
for standards-compliant testing of single and  
3-phase, mode 2 and 3 charging cables with  
simulation of faults in accordance with DIN  
EN 50678 VDE 0701 / DIN EN 50699 VDE  
0702 and per manufacturer’s specifications.  
Some measurements/tests can be per-  
formed with the adapter only, others require  
a compatible test instrument or accessory.  
Fully unroll all measurement cables  
before starting a test/measurement.  
Never perform a test/measurement with  
the measurement cable rolled up.  
When using a test probe with coil cord:  
Grip the tip of the test probe firmly, for  
example if it has been inserted into a jack  
socket. Tensioning at the coil cord may  
otherwise cause the test probe to snap  
back resulting in possible injury.  
Test Adapter  
The following faults can be simulated using  
the test adapter with mains supply to a  
charging cable:  
Interruption of individual conductors  
(undervoltage detection)  
Reversed wires  
Fuses  
The instrument may only be used as long  
PE to phase  
as the fuses are in flawless condition.  
Defective fuses must be replaced.  
Never bridge the fuses. Never put the  
fuses out of operation.  
(Interference voltage in the protective  
conductor due to connection of the  
phase conductor to the protective con-  
ductor)  
Evaluation of the charging cable’s reaction to  
each fault is strictly visual in this case via the  
ICCB indicator lamp or the LEDs at the test  
adapter.  
Test Adapter + Test Instrument  
The following additional measurements/tests  
are possible when a compatible test instru-  
ment (see below) is connected:  
Protective conductor resistance mea-  
surement of the charging cable  
Insulation resistance measurement of the  
charging cable  
RCD tripping test at the ICCB with nomi-  
nal residual current  
(not possible with SECUTEST ST PRO.)  
Measurement of time to trip of the RCD  
at the ICCB  
Guided test sequences (device-specific)  
(not possible with PROFITEST MXTRA  
and PROFITEST MTECH+)  
4 | 22  
   
Documentation  
Compatible test instruments:  
from injury, and to avoid damaging the  
instrument.  
PROFITEST MF XTRA (M534H)  
PROFITEST MF XTRA (LEMONGREEN)  
(M534Y)  
The latest version of these instructions is  
available on our website:  
PROFITEST MF TECH (M534K; M534X)  
PROFITEST MXTRA (M520P; M522P)  
PROFITEST MXTRA IQ (M534M; M534D)  
PROFITEST MTECH+ (M520R; M522R)  
PROFITEST MTECH+ IQ (M534L; M534B)  
PROFITEST PRIME (M506A; M516A)  
PROFITEST PRIME AC (M506C; M516C)  
SECUTEST ST PRO (M707B)  
Errors and Suggestions for Improvement  
These instructions have been prepared with  
utmost care in order to ensure correctness  
and completeness. Unfortunately, errors can  
never be entirely avoided. Continuous  
improvement is part of our quality goal, so  
we always appreciate your comments and  
suggestions.  
Test Adapter + Accessories  
Further measurements/tests can be per-  
formed when special accessories (see  
below) are connected.  
Measurement of protective conductor  
current with current clamp transformer  
METRACLIP 61  
Gender Equality  
For better readability, only the masculine  
form is used in these instructions in a gram-  
matically impartial sense. The feminine and  
diverse forms are of course always implied  
as well.  
Measurement of resistance coding  
with standards-compliant resistance mea-  
suring instrument  
(not possible with SECUTEST ST PRO).  
Safety of the operator, as well as that of the  
test adapter, is only assured when it’s used  
for its intended purpose.  
Trademark Law  
Product designations used in this document  
may be subject to brand law and patent law.  
They are of the property of their respective  
owner.  
2.2  
Use for Other than Intended Purpose  
Using the instrument for any purposes other  
than those described in these instrument  
operating instructions constitutes improper  
use. Use for Other than Intended Purpose  
may lead to unpredictable damage!  
Copyright  
This document is protected by copyright.  
Content modifications, reproduction, dupli-  
cation, processing or translation in any form  
(including excerpts) are only permissible after  
previously obtaining written consent from  
Gossen Metrawatt GmbH. This applies in  
particular to storage and processing in elec-  
tronic systems provided they do not exclu-  
sively serve legitimate internal purposes.  
2.3  
Liability and Guarantee  
Liability and guarantee granted by Gossen  
Metrawatt GmbH complies with the applica-  
ble contractual and mandatory legal regula-  
tions.  
3.2  
Meanings of Symbols in the Operating  
Instructions  
3
Documentation  
LED L1, L2, L3, N or PE at test adapter  
LED on LED off  
3.1  
Information Concerning these  
Instructions  
Read these instructions attentively and care-  
fully. They contain all necessary information  
for safe use of the instrument. Comply with  
them in order to protect yourself and others  
5 | 22  
         
The Instrument  
4
The Instrument  
4.1  
Scope of Delivery  
Please check the scope of delivery for com-  
pleteness and intactness.  
1
Test adapter PROFITEST EMOBILITY  
(M513R) in case  
1
1
Mains power cable  
Set of operating instructions (this docu-  
ment)  
4.2  
Optional Accessories  
Some measurements/tests necessitate  
optional accessories:  
METRACLIP 61 (M311D)  
Digital current clamp  
Adapter Connecting-Cable-16 (Z570B)  
Adapter Connecting-Cable-32 (Z570c)  
standards-compliant resistance measur-  
ing instrument  
Additional adapters and additional informa-  
tion on accessories are available in the data  
sheet of the PROFITEST EMOBILITY.  
For information on the current compatible  
test instruments please refer to the data  
sheets of the respective test instruments.  
You can find the datasheets on our website.  
4.3  
Symbols on the Instrument  
300 V CAT II Maximum permissible voltage  
and measuring category  
between connections and  
ground  
Warning concerning a point of  
danger  
!
(Attention, observe documen-  
tation!)  
Indicates EC conformity  
The device may not be dis-  
posed of with household  
trash.  
For further information see  
“Disposal and Environmental  
Protection” on page 21.  
6 | 22  
       
The Instrument  
4.4  
Connections Overview  
3
4
2
1
10  
5
7
9
8
6
1 N2: Input sockets for 3-phase mains  
connection with the help of a CEE  
adapter (16 A, 32 A with five 4 mm  
6 ICCB-OUT1*: Charging cable output sock-  
ets wired parallel to OUT2 for connecting  
a test instrument for protective conduc-  
tor and insulation measurement  
7 OUT2: Socket (MENNEKES) for connect-  
ing a mode 2 charging cable (vehicle  
end) via a type 2 charging plug  
safety sockets) and mains connection  
fuses F1, F2 and F3 for the 3 phases  
2 N1: Inlet plug for mains connection,  
mains connection fuses F for L and N  
LN  
3 IN2: Socket (MENNEKES) for connecting  
a 3-phase, mode 3 charging cable  
(charging station end) via a type 2  
charging plug  
8 IN3: Input socket in order to be able to  
connect a charging cable with specific  
plug per IEC 62196 or CEE plug  
(charging station end) via adapter  
4 ICCB-IN1*: Input sockets wired parallel to  
connection sockets IN2, IN3 and IN4 for  
connecting a test instrument for protec-  
tive conductor and insulation measure-  
ment  
9 IN4: Earthing contact input socket for  
connecting the supply plug of a single-  
phase, mode 2 charging cable (charging  
station end)  
10  
I
: Loop for measuring protective conduc-  
PE  
5 Rotary selector switches (S1 and S2):  
tor current with a current clamp trans-  
former  
See description on page 3.  
* ICCB = in-cable control box: control box inside the mode 2 charging cable  
7 | 22  
 
The Instrument  
4.5  
Operating Overview  
Top Switch  
Mains connect, mode 2 charging cable  
Status B, C, E  
– for fault simulation  
– for R measurement  
PE  
– for tripping test with I  
nom  
and measurement of time to trip  
Mains disconnect, mode 2/3  
charging cable  
Riso  
– for RISO measurement  
Fault Selection  
Interrupted phase  
Reversed wires  
Bottom Switch  
Orange switch positions: Measurements/tests at single-phase, mode 2/3 cables  
Green switch positions:  
Measurements/tests at 3-phase, mode 2/3 cables  
4.6  
Relevant Standards  
The test adapter has been manufactured  
and tested in accordance with the following  
safety regulations:  
DIN EN  
Electrical safety in low voltage distribution sys-  
61557-16 tems up to 1000 V AC and 1500 V DC –  
Equipment for testing, measuring or monitor-  
ing of protective measures –  
Part 16: Devices for testing the effectiveness  
of protective measures of electrical devices  
and/or electrical medical devices  
DIN EN  
60529  
Degrees of protection provided by enclosures  
(IP code)  
DIN EN IEC Electric vehicle conductive charging systems  
61851-1 Part 1: General requirements  
DIN EN  
Safety requirements for electrical equipment  
61010-1 for measurement, control and laboratory use –  
Part 1: General requirements  
DIN EN IEC Electrical safety in low voltage distribution sys-  
61557-2 tems up to 1000 V AC and 1500 V DC –  
Equipment for testing, measuring or monitor-  
ing of protective measures  
DIN EN  
Electrical safety in low voltage distribution sys-  
61557-4 tems up to 1000 V AC and 1500 V DC–  
Equipment for testing, measuring or monitor-  
ing of protective measures– Part 4: Resis-  
tance of earth conductors, protective conduc-  
tors and equipotential bonding conductors  
8 | 22  
   
The Instrument  
Mechanical Design  
Protection  
4.7  
Technical Data  
Housing: IP40  
Terminals – test outlets  
(protection against  
ingress of solid for-  
eign objects:  
1.0 mm diameter,  
protection against  
ingress of water:  
not protected)  
Earth contact  
IN4:  
1P+N+PE, 0.8 A,  
230 V  
3P+N+PE  
IN2/OUT2:  
0.8 A, 400 V  
Power Supply  
Terminals IP20  
(protection against  
ingress of solid for-  
eign objects:  
12.50 mm diame-  
ter, protection against  
ingress of water:  
not protected)  
per DIN EN 60529 /  
IEC 60529  
Housing (W × H × D):  
Approx. 401 × 307 ×  
173 mm  
(without connector  
cable, with surface  
mount sockets)  
Approx. 4.6 kg  
(incl. connector  
cable)  
Nominal line voltage 230/400 V 50 Hz  
Mains connection  
Single-phase via  
recessed  
power receptacle:  
230 V 1P+N+PE 16 A  
or  
3-phase via ISO  
adapter:  
230/400 V 3P+N+PE  
16 A  
Dimensions  
Weight  
Throughput rating  
Power consumption  
Earth contact: 20 VA  
CEE: 60 VA  
Earth contact: < 3 VA  
CEE: < 6 VA  
Electrical Safety  
Measuring category 300 V CAT II  
Pollution degree  
Fuse links  
2
Supply network:  
Single-phase (N1):  
FLN: 2 × F0,8A/  
250V, 5 × 20 mm  
3-phase (N2):  
F1, F2 and F3:  
3 ea. F0.8A/500V,  
6.3 x 32 mm  
4.8  
Characteristic Values  
Measurement/test with METRACLIP 61 current  
clamp:  
Protective conductor Measuring range:  
current measurement 0 mA … 30 mA AC  
Measurements/tests with compatible test instru-  
ment:  
Ambient Conditions  
Operating temperature –5 … +50 C  
Storage temperature –20 … +60 C  
Protective conductor  
measurement  
see technical data on  
R
function of the  
LO  
Relative humidity  
Max. 75 %, no con-  
densation allowed  
test instrument  
Insulation  
measurement  
see technical data on  
R
function of the  
ISO  
test instrument  
9 | 22  
   
Initial Startup  
5
Initial Startup  
See the connections overview for all con-  
nection variants on page 7.  
5.2  
Testing the LEDs  
Single-Phase Mains Connection  
Insert the earthing contact plug into the  
earthing contact mains outlet, remove it,  
rotate it 180° and insert it again.  
5.1  
Mains Connection  
The test adapter must be connected to the  
mains for fault simulation, as well as for indi-  
cation by means of the phase LEDs.  
When connected with correct polarity, only  
the ICCB-IN1 L1 LED may light up. When con-  
nected with incorrect polarity or if rotated  
180°, the ICCB-IN1 L1 and ICCB-IN1 PE LEDs  
light up simultaneously.  
Attention!  
!
!
Only one mains power cable (single  
or 3-phase) may be connected to the  
test adapter at any given time.  
3-Phase Mains Connection  
Connect socket N2 to the 3-phase 16 A  
mains with the help of the Z570B  
adapter (or the Z570C adapter for 3-  
phase 32 A connection).  
Attention!  
Despite isolation of mains connec-  
tions N1 and N2, do not touch open  
inlet plug N1 in the case of CEE con-  
nection at N2.  
The ICCB-IN1 L1, L2 and L3 LEDs must light  
up.  
5.3  
Connecting the Mode 2/3 Charging  
Cable  
Connect the test adapter to the mains  
via the single or 3-phase mains power  
cable (with the help of the Z570B  
adapter for 3-phase 16 A mains or the  
Z570C adapter for 3-phase 32 A con-  
nection).  
The respective charging cable must be con-  
nected to the test adapter for all measure-  
ments/tests.  
Connecting a Single-Phase, Mode 2 Charging  
Cable  
Refer to the characteristic values on  
page 9 for nominal mains values.  
Insert the power supply plug of the mode  
2 charging cable into earthing contact  
outlet IN4, or into IN3 at the test adapter  
via an adapter.  
Insert the test object’s plug at the vehicle  
end into test socket OUT2.  
Attention!  
!
Single-Phase Mains Connection  
For correct phase connection, the  
earthing contact plug must inserted  
into the mains outlet such that only  
the ICCB-IN1 L1 LED lights up.  
The ICCB-IN1 PE LED lights up as well  
in the case of polarity reversal.  
Connecting a 3-Phase, Mode 2/3 Charging Cable  
Insert the plug at the power supply end  
of the test object into the outlet of the  
adapter connected to IN3, or into IN2  
(mode 3 charging cable).  
Insert the test object’s plug at the vehicle  
end into test socket OUT2.  
10 | 22  
       
Fault Simulation  
6
Fault Simulation  
Mode 2 and 3 charging cables can be tested  
with the test adapter with simulation of faults.  
6.1.2 Simulated Interruption  
Start with the bottom rotary selec-  
tor switch (S2) in the first ON position in  
the orange area for single-phase inter-  
ruption.  
Switch through the positions, one after  
the other, in the clockwise direction (ta-  
ble from top to bottom):  
Attention!  
!
In any case, comply with the manu-  
facturer’s recommendations con-  
cerning measurements/tests to be  
conducted in accordance with  
DGUV, regulation 3.  
Rotary Test  
switch Adapter  
ICCB Action  
The following faults can be simulated for  
mains power supply to a mode 2 charging  
cable:  
LED PE  
LED  
L1  
LED  
Reversed wires  
Interruption of individual conductors  
(undervoltage detection)  
PE to phase (interference voltage in the  
protective conductor due to connection  
of the phase conductor to the protective  
conductor)  
The test object’s  
1
2
3
4
5
6
correct reaction to  
the simulated fault  
can be found in  
the manufac-  
turer’s operating  
instructions.  
The evaluation of the test object’s reaction to  
each respective fault is strictly visual:  
ICCB active or inactive  
(indicator lamp on the ICCB)  
Fault indication by means of LEDs on the  
test adapter  
6.1  
Mode 2 Charging Cable  
(Single-phase)  
6.1.3 Simulation of PE to Phase (PE-U  
)
EXT  
Switch mains power on by  
Set the bottom rotary selector switch  
(S2) to the PE-U position.  
turning the top rotary selector switch (S1)  
to the Status C position.  
EXT  
Interference voltage is switched to PE with  
touch protection.  
The L1 LED and the PE LED light up.  
6.1.1 Simulation of Reversed Wires  
Turn the bottom rotary selector  
switch (S2) to the L1-PE position in  
the orange area for single-phase wire re-  
versal.  
Rotary Test  
switch Adapter  
ICCB Action  
L1  
PE  
LED  
Instead of the L1 LED, the PE LED must light up.  
LED LED  
The test object’s  
8
Rotary Test  
ICCB Action  
correct reaction to  
the simulated fault  
can be found in the  
manufacturer’s op-  
erating instruc-  
tions.  
switch Adapter  
L1  
PE  
LED  
LED LED  
The test object’s  
7
correct reaction to  
the simulated fault  
can be found in the  
manufacturer’s op-  
erating instructions.  
11 | 22  
   
Fault Simulation  
6.2  
Mode 2 (3-phase) Charging Cable  
Switch mains voltage on by turning the top rotary selector switch (S1) to the  
Status C position (system state: vehicle ready for charging).  
6.2.1 Simulated Interruption  
Start with the bottom rotary selector switch (S2) in the first ON position in the green area  
for single-phase interruption.  
Switch through the positions, one after the other, in the clockwise direction (table from  
top to bottom). Conductor interruption using a 3-phase, mode 3 charging cable as an  
example:  
Note  
Tripping performance may deviate from this example in the case of ICCBs from other man-  
ufacturers – adhere to the manufacturer’s test instructions!  
Rotary Test Adapter  
switch  
ICCB Action  
LED  
L1  
LED LED  
N
PE  
LED L2  
L3  
LED LED  
The test object’s correct reaction to the simulated fault can  
1
2
3
4
5
6
7
8
9
be found in the manufacturer’s operating instructions.  
Regarding step 8: The N LED also lights up due to star con-  
nection, although N is interrupted!  
10  
6.2.2 Simulation of PE to Phase (PE-U  
)
EXT  
Set the bottom rotary selector switch (S2) to the PE-U position.  
EXT  
ICCB-IN1 Action  
LED  
Rotary switch Test Adapter  
L1  
LED LED  
N
PE  
LED L2  
L3  
LED LED  
The test object’s correct reaction to the simu-  
lated fault can be found in the manufacturer’s  
operating instructions.  
11  
12 | 22  
 
Measurements/Tests with Test Instruments  
PROFITEST MF XTRA, PROFITEST MF TECH,  
PROFITEST MXTRA, PROFITEST MXTRA IQ,  
PROFITEST MTECH+, PROFITEST MTECH+ IQ:  
2-pole measuring adapter  
PROFITEST PRIME, PROFITEST PRIME AC:  
probes L(1) and PE(3)  
SECUTEST ST PRO: 2-pole measurement  
with test probes P1 and P2 (P1-P2)  
Switch the test object’s ICCB on.  
Perform the measurement as described  
in the operating instructions for the re-  
spective test instrument.  
7
Measurements/Tests with Test  
Instruments  
7.1  
General Information  
When connecting a compatible test instru-  
ment, additional measurements/tests are  
possible which are described in this chapter.  
Important information for the performance of  
measurements/tests  
Attention!  
!
When turning the switch, pay atten-  
tion to the exact snap-in position and  
the corresponding optical signals of  
the LEDs, as described in these oper-  
ating instructions.  
7.2.2 Mode 3 Charging Cable  
The rotary selector switch must be set to  
the R position.  
ISO  
Connect the test object as described in  
The test adapter must remain connected to  
the mains for protective conductor measure-  
Connect the test object to the ICCB-IN1  
PE and ICCB-OUT1 PE sockets:  
PROFITEST MF XTRA, PROFITEST MF TECH,  
PROFITEST MXTRA, PROFITEST MXTRA IQ,  
PROFITEST MTECH+, PROFITEST MTECH+ IQ:  
2-pole measuring adapter  
PROFITEST PRIME, PROFITEST PRIME AC:  
probes L(1) and PE(3)  
SECUTEST ST PRO: 2-pole measurement  
with test probes P1 and P2 (P1-P2)  
Perform the measurement as described  
in the operating instructions for the re-  
spective test instrument.  
ments (  
R
) at mode 2 charging cables by  
PE  
means of the respective test instrument.  
The test adapter can remain connected  
to the mains for protective conductor  
measurements conducted at test  
objects whose protective conductor is  
not switched, as well as for insulation  
measurements (Riso) with the respective  
test instrument.  
7.2  
Measurement of Protective Conduc-  
tor Resistance (R )  
LO  
7.2.1 Mode 2 Charging Cable  
7.3  
Measuring Insulation Resistance (R ISO)  
As opposed to the usual default setting for  
low-resistance measurements, the device  
under test does not have to be disconnected  
from all sources of voltage in this case. Both  
rotary selector switches (S1 and S2) have to be  
set to the ON and the Status C positions, in  
order to be able to activate the test object and  
switch the protective conductor through. Prior  
to a protective conductor resistance measure-  
ment, an offset measurement must be carried  
out (the offset value of the test adapter must  
be taken into account during the test).  
Measurement is carried out between ICCB-  
OUT1-PE and OUT2-PE.  
Attention!  
!
Insulation resistance may only be  
measured at voltage-free devices!  
The upper-hand rotary selector  
switch must be set to the mains dis-  
connect position: R  
INS  
Connect the test instrument to the ICCB-  
IN1 socket for the mains-end of the cable  
and the ICCB-OUT1 socket for the output  
end of the cable of the device under test:  
PROFITEST MF XTRA, PROFITEST MF TECH,  
PROFITEST MXTRA, PROFITEST MXTRA IQ,  
PROFITEST MTECH+, PROFITEST MTECH+ IQ:  
2-pole measuring adapter  
Connect the test object as described in  
Connect the test object to the ICCB-IN1  
PE and ICCB-OUT1 PE sockets:  
PROFITEST PRIME, PROFITEST PRIME AC:  
probes L(1) and PE(3)  
13 | 22  
       
Measurements/Tests with Test Instruments  
SECUTEST ST PRO: 2-pole measurement  
We recommend the following procedure in  
order to prevent an upstream RCD (selective  
RCD as well) from tripping, and to ensure  
that actual tripping current or correct time to  
trip is measured:  
with test probes P1 and P2 (P1-P2)  
Connect the test object as described in  
Mode 2 Charging Cable:  
Connect the respective test instrument  
with pole (L1) to the voltage conducting  
pole depending on the earthing contact  
plug wiring of charging cable L1, N (ICCB-  
OUT1), and with pole (PE) to N (ICCB-IN1).  
(observe limit values and manufacturer’s  
specifications):  
Connect the test instrument to the ICCB-  
IN1/ICCB-OUT1 L1 or N and PE sockets,  
one after the other:  
PROFITEST MF XTRA, PROFITEST MF TECH,  
PROFITEST MXTRA, PROFITEST MXTRA IQ,  
PROFITEST MTECH+, PROFITEST MTECH+ IQ:  
2-pole measuring adapter  
PROFITEST PRIME, PROFITEST PRIME AC:  
probes L(1) and PE(3)  
SECUTEST ST PRO: 2-pole measurement  
with test probes P1 and P2 (P1-P2)  
Perform the measurement as described  
in the operating instructions for the re-  
spective test instrument.  
PROFITEST MF XTRA, PROFITEST MF TECH,  
PROFITEST MXTRA, PROFITEST MXTRA IQ,  
PROFITEST MTECH+, PROFITEST MTECH+ IQ:  
2-pole measuring adapter  
PROFITEST PRIME, PROFITEST PRIME AC:  
probes L(1) and PE(3)  
Refer to the operating instructions for the  
respective test instrument regarding per-  
formance of the measurement.  
Mode 3 Charging Cable  
Connect the respective test instrument  
to the ICCB-IN1/ICCB-OUT1 L1, L2, L3, N  
and PE sockets, one after the other:  
PROFITEST MF XTRA, PROFITEST MF TECH,  
PROFITEST MXTRA, PROFITEST MXTRA IQ,  
PROFITEST MTECH+, PROFITEST MTECH+ IQ:  
2-pole measuring adapter  
PROFITEST PRIME, PROFITEST PRIME AC:  
probes L(1) and PE(3)  
SECUTEST ST PRO: 2-pole measurement  
with test probes P1 and P2 (P1-P2)  
7.4  
Tripping Test with Nominal Residual  
Current and Measurement of Time to  
Trip at Mode 2 Charging Cable  
(not possible with SECUTEST ST PRO)  
Note  
If you operate your test adapter at an  
electrical system with a 30 mA  
breaker, the mains RCD may be  
tripped during the tripping test (trip-  
ping current or time to trip) instead of  
the charging cable’s ICCB. In this  
case the ICCB-IN1 L1 LED goes out.  
14 | 22  
 
Protective Conductor Current Measurement (IPE) at Mode 2 Charging Cable  
7.5  
Guided Test Sequences  
8
Protective Conductor Current  
Measurement (IPE) at Mode 2  
Charging Cable  
(not possible with PROFITEST MXTRA  
and PROFITEST MTECH+)  
Test sequences are available for the test  
instruments. The test instrument runs  
through all test steps semi-automatically.  
Each test step is evaluated and assessed by  
the user (go/no-go) for later documentation.  
It depends on the respective test instrument  
whether or not and which test sequences  
are available.  
The test sequences are provided free of  
charge for download from the myGMC por-  
tal, however, it is the responsibility of the  
user to import them into the test instrument.  
Protective conductor current or bias current  
may result in premature tripping of RCDs  
used in ICCBs.  
For this reason, the protective conductor is  
led out of the housing as a loop between  
surface mount sockets 4 and 6. This makes  
it possible to measure any protective con-  
ductor current with the help of a current  
clamp transformer.  
You therefore require an additional current  
clamp transformer (e.g. METRACLIP 61).  
Connect the test object as described in  
Switch mains voltage on by turning the  
top rotary selector switch (S1) to the Sta-  
tus C position (system state: vehicle  
ready for charging).  
Create a myGMC account and register  
your test instrument (both free of  
charge):  
Turn the bottom rotary selector switch  
(S2) to the first ON position in the orange  
area for single-phase interruption.  
Close the jaws of the current clamp  
transformer around the external protec-  
Download the test sequence(s).  
Transfer the test sequence(s) to your test  
instrument. Complete information can be  
found in the product documentation for  
the respective test instrument.  
Perform the test sequence with the test  
instrument and the test adapter.  
Detailed information on the test se-  
quences can be found in the product  
documentation of the respective test in-  
strument.  
tive conductor loop I .  
PE  
Read the measured value for protective  
conductor current at the current clamp  
transformer.  
Protective conductor current should not  
exceed 3.5 mA.  
15 | 22  
   
Measuring RC (Testing of Resistance Coding) (not possible with SECUTEST ST PRO)  
9
Measuring RC  
10  
Storage and Transport  
(Testing of Resistance Coding)  
(not possible with SECUTEST ST  
PRO)  
Attention!  
Improper Storage  
!
Damage to the product and measur-  
ing error due to environmental influ-  
ences  
Store the instrument in a protected  
location and only within the limits of  
permissible ambient conditions. The  
respective ambient conditions (tem-  
perature, humidity etc.) can be found  
Testing of specified resistance value R for  
maximum current-carrying capacity of the  
mode 2/3 charging cable per  
DIN EN 61851-1 (VDE 0122-1)  
Additionally, you require a standards-compli-  
ant resistance measuring instrument.  
C
Connect the test object as described in  
Measure resistance between the PP and  
PE sockets at ICCB-OUT1 with a stan-  
dards-compliant measuring instrument.  
Compare the measured resistance value  
with the value indicated in the table  
below (Table B3 off DIN EN 61851-1).  
Note  
If the test adapter has not been used  
for a long period of time, the switches  
may demonstrate increased contact  
resistance depending upon storage  
conditions.  
If this is the case, actuate the  
switches several times.  
Note  
Observe the maximum current-carry-  
ing capacity specified on the cable.  
Vehicle state  
Function Test Result  
Attention!  
!
Testing of resis- Resistance 13 A charging cable,  
Improper Transport  
tance coding for measure-  
vehicle inlets and ment  
1.5 k  
20 A charging cable, 680  
Damage to the product and measur-  
ing error  
Transport the instrument only within  
the limits of permissible ambient con-  
ditions (temperature, humidity etc.).  
vehicle connec- with multi-  
tors per  
meter or  
test instru-  
ment  
32 A charging cable, 220  
IEC 61851  
Table B.3  
63 A charging cable, 100  
16 | 22  
   
Maintenance  
11.2.1 Testing of Protective Conductor Resis-  
tance R  
11  
Maintenance  
PE  
11.1 Cleaning of Housing  
Test Adapter Contacting, Single-Phase (230 V)  
Protective conductor resistance is tested  
between the PE contact at the mains plug  
(inlet plug N1) and PE contacts ICCB IN1  
through IN4.  
Attention!  
Life endangering due to electric  
shock!  
!
The test adapter is operated with  
electric power. Therefore there is a  
general risk of electric shock which  
can be fatal or cause severe injury.  
– The test adapter must be voltage-  
free both before and during cleaning.  
Switch the test adapter off and dis-  
connect it from the mains power  
supply.  
– The test adapter may not be con-  
nected to a measuring circuit before  
and during cleaning.  
– Never immerse the test adapter in  
water or other liquids.  
Testing the Single-Phase Connection  
Turn the top rotary selector switch (S1) at  
the test adapter to the Status C position  
(system state: vehicle ready for  
charging).  
With the exclusion of the exceptions speci-  
fied in the following table, a protective con-  
ductor resistance R of less than 2 is  
PE  
permissible. This is due to the design of the  
test adapter.  
Rotary Test  
switch Adapter  
ICCB Protective  
Conductor Re-  
sistance R PE,  
– Never touch the test adapter to  
with wet or moist hands.  
L1  
PE  
LED Permissible  
LED LED  
No special maintenance is required. Keep  
outside surfaces clean and dry.  
1
2
3
4
5
6
< 2 W  
< 2 W  
< 2 W  
< 2 W  
< 2 W  
> 30 MW  
Attention!  
!
Unsuitable cleaning agents such as  
aggressive or abrasive cleansers re-  
sult in damage to the test adapter.  
– Use a cloth for cleaning which has  
been slightly dampened with water.  
– Avoid the use of cleansers, abra-  
sives or solvents.  
11.2 Technical Safety Inspections  
Testing per DGUV Regulation 3  
Subject your test adapter to technical safety  
inspections at regular intervals.  
The test adapter is designed in accordance  
with IEC 61010 as a protection category I  
and II test instrument.  
7
8
> 30 MW  
> 30 MW  
Testing of the protective conductor, insula-  
tion resistance and touch current is  
described in the following subsections.  
Test Adapter Contacting, 3-Phase (400 V)  
Protective conductor resistance is tested  
between the PE contact at mains connec-  
tion N2 and the PE contact at ICCB-IN1  
(parallel to IN2, IN3 and IN4).  
17 | 22  
     
Maintenance  
Testing the 3-Phase Connection  
at the power supply end of the test  
object, ICCB-IN1  
at the output end of the test object, ICCB-  
OUT1  
Turn the top rotary selector switch (S1) at  
the test adapter to the Status C position  
(system state: vehicle ready for  
charging).  
in each case against PE.  
The usual limit values apply.  
Turn the bottom rotary selector switch  
(S2) to each switch position*.  
11.2.3 Touch Current Measurement  
With the exclusion of the exceptions specified  
in the following table, a protective conductor  
Touch current measurement is conducted at  
the screw connections of the IN sockets  
(IN2, IN3 and IN4) with the standard limit val-  
resistance R of less than 2  
is permissible.  
PE  
This is due to the design of the test adapter.  
ues (I < 0,5 mA).  
B
Rotary  
switch  
Protective Conductor Resis-  
tance R PE,  
11.3 Fuse Replacement  
The test adapter may only be operated with  
fuses in flawless condition. Damaged or  
defective fuses must be replaced immedi-  
ately.  
All fuses for neutral and phase conductors  
are accessible from the outside.  
Permissible  
< 2 W  
1
2
< 2 W  
3
< 2 W  
Attention!  
Life endangering due to electric  
shock!  
!
4
< 2 W  
The test adapter is operated with  
electric power. Therefore there is a  
general risk of electric shock which  
can be fatal or cause severe injury.  
– The test adapter must be voltage-  
free both before and during changing  
the fuse. Switch the test adapter off  
and disconnect it from the mains  
power supply.  
5
< 2 W  
6
< 2 W  
7
< 2 W  
8
< 2 W  
– The test adapter may not be con-  
nected to a measuring circuit before  
and during cleaning.  
9
< 2 W  
10  
11  
> 30 MW  
> 30 MW  
Attention!  
!
Only install fuses of a type that con-  
forms to the details specified in the  
Technical Data, see page 9.  
No other fuses may not be used in  
the test instrument.  
The PE connection between OUT1 and  
OUT2 must also be tested (< 2 ).  
11.2.2 Testing Insulation Resistance  
Testing is conducted in the respective switch  
positions for Status B, C or E and in the  
green ON position for 3-phase testing at  
short-circuited L-N or L1-L2-L3-N contacts  
(for earthing contact and CEE respectively)  
at mains connections N1 and N2  
18 | 22  
 
Repair  
F1, F2 and F3 (3-phase mains supply power sys-  
tem N2)  
Required materials:  
12  
Repair  
If your instrument requires repair, please  
contact our service department, see “Con-  
A slotted screwdriver for opening the  
fuse compartment  
Replacement fuses in accordance with  
the details specified in the Technical Data  
Note  
Loss of warranty and guarantee  
claims!  
Unauthorized modification of the tes-  
ter is prohibited. This also includes  
opening the meter.  
If it can be ascertained that the tester  
has been opened by unauthorized  
personnel, no guarantee claims can  
be honored by the manufacturer with  
regard to personal safety, measuring  
accuracy, compliance with applicable  
safety measures or any consequen-  
tial damages.  
Open the respective fuse compartment.  
Insert the slotted screwdriver into the slot  
on the fuse compartment and turn it  
counterclockwise.  
Pull out the fuse cover as soon as it pro-  
trudes slightly from the housing.  
Remove the defective fuse from the fuse  
cover.  
Insert the new fuse into the fuse cover.  
Insert the fuse cover with the fuse back  
into the fuse compartment. The two tabs  
on the fuse cover must be inserted into  
the recesses in the housing.  
Insert the slotted screwdriver into the slot  
on the fuse compartment and turn it  
clockwise until the fuse compartment is  
closed.  
If the guarantee seal is damaged or  
removed, all guarantee claims are  
rendered null and void.  
The device may only be repaired or opened  
by authorized, qualified personnel who are  
familiar with the associated dangers.  
Original replacement parts may only be  
installed by authorized, qualified personnel.  
The instrument may not be placed back into  
operation until troubleshooting and repair  
have been performed, and calibration and  
dielectric strength have been tested and  
approved at our factory or by an authorized  
service center.  
FLN (mains power supply system single-phase N1)  
You require replacement fuses in accor-  
dance with the details specified in the Tech-  
nical Data on page 9.  
Open the fuse compartment. Press to-  
gether the plastic wings of the fuse link  
to this end and pull it up slightly.  
As soon as the fuse link protrudes a little  
bit, pull it out.  
Remove the defective fuses from the  
fuse link.  
Insert the new fuses into the fuse link.  
Insert the fuse link with the fuses into the  
fuse compartment. The tab on the fuse  
cover must be inserted into the recesses  
in the housing.  
Press down the fuse link until it snaps  
into place.  
19 | 22  
 
Contact, Support and Service  
13  
Contact, Support and Service  
14  
CE Declaration  
Gossen Metrawatt GmbH can be reached  
directly and simply – we have a single num-  
The instrument fulfills all requirements of  
applicable EU directives and national regula-  
ber for everything! Whether you require sup- tions. We confirm this with the CE mark. The  
port or training, or have an individual inquiry,  
we can answer all of your questions here:  
CE declaration can be found on our website:  
+49-911-8602-0  
Monday to  
Thursday:  
Friday:  
8 a.m. to 4 p.m.  
8 a.m. to 2 p.m.  
Or contact us by e-mail at:  
Do you prefer support by e-mail?  
Measuring and Test Technology:  
Industrial Measuring Technology:  
Please contact GMC-I Service GmbH for  
repairs, replacement parts and calibration*:  
+49-911-817718-0  
Beuthener Str. 41  
90471 Nürnberg  
Germany  
*
DAkkS calibration laboratory per DIN EN ISO/IEC 17025  
accredited by the Deutsche Akkreditierungsstelle GmbH under  
reference number D-K-15080-01-01.  
20 | 22  
   
Disposal and Environmental Protection  
limit value for the respective metal.  
15  
Disposal and Environmental  
Protection  
Please observe the owner’s or end user’s  
responsibility with regard to deleting per-  
sonal data, as well as any other sensitive  
data, from old devices before disposal.  
Old devices, electrical or electronic accesso-  
ries and waste batteries (including recharge-  
able batteries) used in Germany can be  
returned free of charge to Gossen Metrawatt  
GmbH or the service provider responsible  
for their disposal in compliance with applica-  
ble regulations, in particular laws concerning  
packaging and hazardous goods. Waste  
batteries must be handed over in discharged  
state and/or with appropriate precautionary  
measures against short-circuiting. Further  
information regarding returns can be found  
on our website.  
Proper disposal makes an important contri-  
bution to the protection of our environment  
and the conservation of natural resources.  
Attention!  
!
Environmental Damage  
Improper disposal results in environ-  
mental damage.  
Follow the instructions concerning  
return and disposal included in this  
section.  
The following comments refer specifically to  
the legal situation in the Federal Republic of  
Germany. Owners or end users who are  
subject to other regulations must comply  
with the respective local requirements and  
implement them correctly on site. Further  
information can be obtained, for example,  
from the responsible authorities or the local  
distributor.  
Packaging Materials  
We recommend retaining the respective  
packaging materials for the case that you  
might require servicing or calibration in the  
future.  
Waste Electrical Equipment, Electrical or Elec-  
tronic Accessories and Waste Batteries (including  
rechargeable batteries)  
Electrical equipment and batteries (including  
rechargeable batteries) contain valuable raw  
materials that can be recycled, as well as  
hazardous substances which can cause  
serious harm to human health and the envi-  
ronment, and they must be recycled and  
disposed of correctly.  
Attention!  
!
Danger of Asphyxiation Resulting  
from Foils and Other Packaging Ma-  
terials.  
Children and other vulnerable per-  
sons may suffocate if they wrap  
themselves in packaging materials,  
or their components or foils, or if they  
pull them over their heads or swallow  
them. Keep packaging materials, as  
well as their components and foils,  
out of the reach of babies, children  
and other vulnerable persons.  
The symbol at the left depicting a  
crossed-out garbage can on  
wheels refers to the legal obligation  
of the owner or end user (German  
electrical and electronic equipment act Elek-  
troG and German battery act BattG) not to  
dispose of used electrical equipment and  
batteries with unsorted municipal waste  
(“household trash”). Waste batteries must be  
removed from the old device (where possi-  
ble) without destroying them and the old  
device and the waste batteries must be dis-  
posed of separately. The battery type and its  
chemical composition are indicated on the  
battery’s labelling. If the abbreviations “Pb”  
for lead, “Cd” for cadmium or “Hg” for mer-  
cury are included, the battery exceeds the  
In accordance with German packaging law  
(VerpackG), the user is obligated to correctly  
dispose of packaging and its components  
separately, and not together with unsorted  
municipal waste (“household trash”).  
Private end consumers can dispose of pack-  
aging free of charge at the responsible col-  
lection point. Packaging which is not subject  
to so-called system participation is returned  
to the appointed service provider. Further  
information regarding returns can be found  
on our website.  
21 | 22  
 
©Gossen Metrawatt GmbH  
Prepared in Germany • Subject to change / errors excepted • A pdf version is available on the Internet  
All trademarks, registered trademarks, logos, product names and company names are the property of their respective  
owners.  
Gossen Metrawatt GmbH  
Südwestpark 15  
90449 Nürnberg  
Germany  
Phone +49 911 8602-0  
Fax +49 911 8602-669  

Gossen MetraWatt A3 32 A3 63 A3 16 Operating Instructions User Manual
Gossen MetraWatt AKKUPACK Z502H Supplement User Manual
Gossen MetraWatt CP 30 1100 1800 330 Operating Instructions User Manual
Gossen MetraWatt DUSPOL expert 1000 M611e Operating User Manual
Gossen MetraWatt KE7010 Quick Start User Manual
Gossen MetraWatt M273A Operating Instructions User Manual
Gossen MetraWatt METRAHIT IM E DRIVE FUSO Short Form Operating Instructions User Manual
Gossen MetraWatt PROFITEST H+E EASY CHECK Operating Instructions User Manual
Gossen MetraWatt PROFITEST PRIME Short Form Operating Instructions User Manual
Gossen MetraWatt U1601 U1602 U1603 Operating Instructions User Manual