Gossen MetraWatt PROFITEST EMOBILITY Operating Instructions User Manual |
Operating Instructions
PROFITESTEMOBILITY
Adapter for Standards-Compliant Testing of Single and 3-Phase,
Mode 2 and 3 Charging Cables
3-349-981-15
7/2.24
Safety Instructions
1
Safety Instructions
Read and follow these instructions carefully
and completely in order to ensure safe and
proper use.
from operation permanently and secure
it against inadvertent use.
•
•
If there are any signs of interior damage
to the instrument or accessories (e.g.
loose parts in the housing), permanently
remove the instrument/accessories from
operation and secure them against inad-
vertent use.
Route cables in an orderly fashion, e.g.
the mains power cable. Loose, disor-
derly cables result in unnecessary dan-
ger of tripping and falling.
The instrument may only be used for the
measurements/tests described in the
documentation for the instrument.
The instrument may only be connected to
electrical systems with a maximum of 230/
400 V which comply with applicable
safety regulations (e.g. IEC 60346, VDE
0100) and are protected with a fuse or cir-
cuit breaker with a maximum rating of
16 A.
The test adapter may only be used for
testing mode 2 and 3 charging cables.
No power consuming devices may be
connected to any of the sockets.
Measurements within electrical systems
are prohibited.
Read the operating instructions for the
respective test instrument as well, in particu-
lar the sections concerning the R
ISO
PE, R
and time-to-trip measurements, as well as
the tripping test.
General
•
Measurements/tests may only be per-
formed by a qualified electrician, or
under the supervision and direction of a
qualified electrician. The user must be
instructed by a qualified electrician con-
cerning performance and evaluation of
the tests/measurements.
•
•
•
•
Observe and comply with all safety regu-
lations which are applicable for your
work environment.
Wear suitable and appropriate personal
protective equipment (PPE) whenever
working with the instrument.
•
•
•
•
Handling
•
The instruments and accessories of
Gossen Metrawatt GmbH are designed
such as to ensure optimum compatibility
with the Gossen Metrawatt GmbH prod-
ucts that are expressly provided for
them. Unless otherwise expressly con-
firmed in writing by Gossen Metrawatt
GmbH, they are not intended and suited
for use with other products.
Observe the manufacturer’s specifica-
tions for DUTs.
Operating Conditions
•
Do not use the instrument and its acces-
sories after long periods of storage under
unfavorable conditions (e.g. humidity,
dust or extreme temperature).
Do not use the instrument and its acces-
sories after extraordinary stressing due
to transport.
The instrument must not be exposed to
direct sunlight.
Only use the instrument and its accesso-
ries within the limits of the specified tech-
nical data and conditions (ambient con-
ditions, IP protection code, measuring
category etc.).
•
Use the instrument in undamaged condi-
tion only.
Inspect the instrument before use. Pay
particular attention to damage, inter-
rupted insulation or kinked cables.
Also test the LEDs: if the LEDs are
defective, voltage at the ICCB-IN1 sock-
ets is no longer indicated.
•
•
•
•
•
If the instrument doesn’t function flaw-
lessly, permanently remove it from oper-
ation and secure it against inadvertent
use.
If the instrument is damaged during use,
for example if it’s dropped, remove it
3 | 22
Application
Measurement Cables and Establishing Contact
2
Application
•
•
•
Plugging in the measurement cables
must not necessitate any undue force.
Never touch conductive ends (e.g. of
test probes).
Do not touch the insulation measuring
instrument’s test probes during insula-
tion resistance measurements!
2.1
Intended Use /
Use for Intended Purpose
The PROFITEST EMOBILITY is an adapter
for standards-compliant testing of single and
3-phase, mode 2 and 3 charging cables with
simulation of faults in accordance with DIN
EN 50678 VDE 0701 / DIN EN 50699 VDE
0702 and per manufacturer’s specifications.
Some measurements/tests can be per-
formed with the adapter only, others require
a compatible test instrument or accessory.
•
•
Fully unroll all measurement cables
before starting a test/measurement.
Never perform a test/measurement with
the measurement cable rolled up.
When using a test probe with coil cord:
Grip the tip of the test probe firmly, for
example if it has been inserted into a jack
socket. Tensioning at the coil cord may
otherwise cause the test probe to snap
back resulting in possible injury.
Test Adapter
The following faults can be simulated using
the test adapter with mains supply to a
charging cable:
•
Interruption of individual conductors
(undervoltage detection)
Reversed wires
Fuses
•
•
•
The instrument may only be used as long
PE to phase
as the fuses are in flawless condition.
Defective fuses must be replaced.
Never bridge the fuses. Never put the
fuses out of operation.
(Interference voltage in the protective
conductor due to connection of the
phase conductor to the protective con-
ductor)
•
Evaluation of the charging cable’s reaction to
each fault is strictly visual in this case via the
ICCB indicator lamp or the LEDs at the test
adapter.
Test Adapter + Test Instrument
The following additional measurements/tests
are possible when a compatible test instru-
ment (see below) is connected:
•
•
•
Protective conductor resistance mea-
surement of the charging cable
Insulation resistance measurement of the
charging cable
RCD tripping test at the ICCB with nomi-
nal residual current
(not possible with SECUTEST ST PRO.)
Measurement of time to trip of the RCD
at the ICCB
Guided test sequences (device-specific)
(not possible with PROFITEST MXTRA
and PROFITEST MTECH+)
•
•
4 | 22
Documentation
Compatible test instruments:
from injury, and to avoid damaging the
instrument.
•
•
PROFITEST MF XTRA (M534H)
PROFITEST MF XTRA (LEMONGREEN)
(M534Y)
The latest version of these instructions is
available on our website:
•
•
•
•
•
•
•
•
PROFITEST MF TECH (M534K; M534X)
PROFITEST MXTRA (M520P; M522P)
PROFITEST MXTRA IQ (M534M; M534D)
PROFITEST MTECH+ (M520R; M522R)
PROFITEST MTECH+ IQ (M534L; M534B)
PROFITEST PRIME (M506A; M516A)
PROFITEST PRIME AC (M506C; M516C)
SECUTEST ST PRO (M707B)
Errors and Suggestions for Improvement
These instructions have been prepared with
utmost care in order to ensure correctness
and completeness. Unfortunately, errors can
never be entirely avoided. Continuous
improvement is part of our quality goal, so
we always appreciate your comments and
suggestions.
Test Adapter + Accessories
Further measurements/tests can be per-
formed when special accessories (see
below) are connected.
•
Measurement of protective conductor
current with current clamp transformer
METRACLIP 61
Gender Equality
For better readability, only the masculine
form is used in these instructions in a gram-
matically impartial sense. The feminine and
diverse forms are of course always implied
as well.
•
Measurement of resistance coding
with standards-compliant resistance mea-
suring instrument
(not possible with SECUTEST ST PRO).
Safety of the operator, as well as that of the
test adapter, is only assured when it’s used
for its intended purpose.
Trademark Law
Product designations used in this document
may be subject to brand law and patent law.
They are of the property of their respective
owner.
2.2
Use for Other than Intended Purpose
Using the instrument for any purposes other
than those described in these instrument
operating instructions constitutes improper
use. Use for Other than Intended Purpose
may lead to unpredictable damage!
Copyright
This document is protected by copyright.
Content modifications, reproduction, dupli-
cation, processing or translation in any form
(including excerpts) are only permissible after
previously obtaining written consent from
Gossen Metrawatt GmbH. This applies in
particular to storage and processing in elec-
tronic systems provided they do not exclu-
sively serve legitimate internal purposes.
2.3
Liability and Guarantee
Liability and guarantee granted by Gossen
Metrawatt GmbH complies with the applica-
ble contractual and mandatory legal regula-
tions.
3.2
Meanings of Symbols in the Operating
Instructions
3
Documentation
LED L1, L2, L3, N or PE at test adapter
LED on LED off
3.1
Information Concerning these
Instructions
Read these instructions attentively and care-
fully. They contain all necessary information
for safe use of the instrument. Comply with
them in order to protect yourself and others
5 | 22
The Instrument
4
The Instrument
4.1
Scope of Delivery
Please check the scope of delivery for com-
pleteness and intactness.
1
Test adapter PROFITEST EMOBILITY
(M513R) in case
1
1
Mains power cable
Set of operating instructions (this docu-
ment)
4.2
Optional Accessories
Some measurements/tests necessitate
optional accessories:
•
METRACLIP 61 (M311D)
Digital current clamp
•
•
•
Adapter Connecting-Cable-16 (Z570B)
Adapter Connecting-Cable-32 (Z570c)
standards-compliant resistance measur-
ing instrument
Additional adapters and additional informa-
tion on accessories are available in the data
sheet of the PROFITEST EMOBILITY.
For information on the current compatible
test instruments please refer to the data
sheets of the respective test instruments.
You can find the datasheets on our website.
4.3
Symbols on the Instrument
300 V CAT II Maximum permissible voltage
and measuring category
between connections and
ground
Warning concerning a point of
danger
!
(Attention, observe documen-
tation!)
Indicates EC conformity
The device may not be dis-
posed of with household
trash.
For further information see
“Disposal and Environmental
Protection” on page 21.
6 | 22
The Instrument
4.4
Connections Overview
3
4
2
1
10
5
7
9
8
6
1 N2: Input sockets for 3-phase mains
connection with the help of a CEE
adapter (16 A, 32 A with five 4 mm
6 ICCB-OUT1*: Charging cable output sock-
ets wired parallel to OUT2 for connecting
a test instrument for protective conduc-
tor and insulation measurement
7 OUT2: Socket (MENNEKES) for connect-
ing a mode 2 charging cable (vehicle
end) via a type 2 charging plug
safety sockets) and mains connection
fuses F1, F2 and F3 for the 3 phases
2 N1: Inlet plug for mains connection,
mains connection fuses F for L and N
LN
3 IN2: Socket (MENNEKES) for connecting
a 3-phase, mode 3 charging cable
(charging station end) via a type 2
charging plug
8 IN3: Input socket in order to be able to
connect a charging cable with specific
plug per IEC 62196 or CEE plug
(charging station end) via adapter
4 ICCB-IN1*: Input sockets wired parallel to
connection sockets IN2, IN3 and IN4 for
connecting a test instrument for protec-
tive conductor and insulation measure-
ment
9 IN4: Earthing contact input socket for
connecting the supply plug of a single-
phase, mode 2 charging cable (charging
station end)
10
I
: Loop for measuring protective conduc-
PE
5 Rotary selector switches (S1 and S2):
tor current with a current clamp trans-
former
See description on page 3.
* ICCB = in-cable control box: control box inside the mode 2 charging cable
7 | 22
The Instrument
4.5
Operating Overview
Top Switch
Mains connect, mode 2 charging cable
Status B, C, E
– for fault simulation
– for R measurement
PE
– for tripping test with I
nom
and measurement of time to trip
Mains disconnect, mode 2/3
charging cable
Riso
– for RISO measurement
Fault Selection
Interrupted phase
Reversed wires
Bottom Switch
Orange switch positions: Measurements/tests at single-phase, mode 2/3 cables
Green switch positions:
Measurements/tests at 3-phase, mode 2/3 cables
4.6
Relevant Standards
The test adapter has been manufactured
and tested in accordance with the following
safety regulations:
DIN EN
Electrical safety in low voltage distribution sys-
61557-16 tems up to 1000 V AC and 1500 V DC –
Equipment for testing, measuring or monitor-
ing of protective measures –
Part 16: Devices for testing the effectiveness
of protective measures of electrical devices
and/or electrical medical devices
DIN EN
60529
Degrees of protection provided by enclosures
(IP code)
DIN EN IEC Electric vehicle conductive charging systems
61851-1 Part 1: General requirements
DIN EN
Safety requirements for electrical equipment
61010-1 for measurement, control and laboratory use –
Part 1: General requirements
DIN EN IEC Electrical safety in low voltage distribution sys-
61557-2 tems up to 1000 V AC and 1500 V DC –
Equipment for testing, measuring or monitor-
ing of protective measures
DIN EN
Electrical safety in low voltage distribution sys-
61557-4 tems up to 1000 V AC and 1500 V DC–
Equipment for testing, measuring or monitor-
ing of protective measures– Part 4: Resis-
tance of earth conductors, protective conduc-
tors and equipotential bonding conductors
8 | 22
The Instrument
Mechanical Design
Protection
4.7
Technical Data
Housing: IP40
Terminals – test outlets
(protection against
ingress of solid for-
eign objects:
1.0 mm diameter,
protection against
ingress of water:
not protected)
Earth contact
IN4:
1P+N+PE, 0.8 A,
230 V
3P+N+PE
IN2/OUT2:
0.8 A, 400 V
Power Supply
Terminals IP20
(protection against
ingress of solid for-
eign objects:
12.50 mm diame-
ter, protection against
ingress of water:
not protected)
per DIN EN 60529 /
IEC 60529
Housing (W × H × D):
Approx. 401 × 307 ×
173 mm
(without connector
cable, with surface
mount sockets)
Approx. 4.6 kg
(incl. connector
cable)
Nominal line voltage 230/400 V 50 Hz
Mains connection
Single-phase via
recessed
power receptacle:
230 V 1P+N+PE 16 A
or
3-phase via ISO
adapter:
230/400 V 3P+N+PE
16 A
Dimensions
Weight
Throughput rating
Power consumption
Earth contact: 20 VA
CEE: 60 VA
Earth contact: < 3 VA
CEE: < 6 VA
Electrical Safety
Measuring category 300 V CAT II
Pollution degree
Fuse links
2
Supply network:
Single-phase (N1):
FLN: 2 × F0,8A/
250V, 5 × 20 mm
3-phase (N2):
F1, F2 and F3:
3 ea. F0.8A/500V,
6.3 x 32 mm
4.8
Characteristic Values
Measurement/test with METRACLIP 61 current
clamp:
Protective conductor Measuring range:
current measurement 0 mA … 30 mA AC
Measurements/tests with compatible test instru-
ment:
Ambient Conditions
Operating temperature –5 … +50 C
Storage temperature –20 … +60 C
Protective conductor
measurement
see technical data on
R
function of the
LO
Relative humidity
Max. 75 %, no con-
densation allowed
test instrument
Insulation
measurement
see technical data on
R
function of the
ISO
test instrument
9 | 22
Initial Startup
5
Initial Startup
See the connections overview for all con-
5.2
Testing the LEDs
Single-Phase Mains Connection
➭
Insert the earthing contact plug into the
earthing contact mains outlet, remove it,
rotate it 180° and insert it again.
5.1
Mains Connection
The test adapter must be connected to the
mains for fault simulation, as well as for indi-
cation by means of the phase LEDs.
When connected with correct polarity, only
the ICCB-IN1 L1 LED may light up. When con-
nected with incorrect polarity or if rotated
180°, the ICCB-IN1 L1 and ICCB-IN1 PE LEDs
light up simultaneously.
Attention!
!
!
Only one mains power cable (single
or 3-phase) may be connected to the
test adapter at any given time.
3-Phase Mains Connection
➭
Connect socket N2 to the 3-phase 16 A
mains with the help of the Z570B
adapter (or the Z570C adapter for 3-
phase 32 A connection).
Attention!
Despite isolation of mains connec-
tions N1 and N2, do not touch open
inlet plug N1 in the case of CEE con-
nection at N2.
The ICCB-IN1 L1, L2 and L3 LEDs must light
up.
5.3
Connecting the Mode 2/3 Charging
Cable
➭
Connect the test adapter to the mains
via the single or 3-phase mains power
cable (with the help of the Z570B
adapter for 3-phase 16 A mains or the
Z570C adapter for 3-phase 32 A con-
nection).
The respective charging cable must be con-
nected to the test adapter for all measure-
ments/tests.
Connecting a Single-Phase, Mode 2 Charging
Cable
➭
Refer to the characteristic values on
page 9 for nominal mains values.
Insert the power supply plug of the mode
2 charging cable into earthing contact
outlet IN4, or into IN3 at the test adapter
via an adapter.
Insert the test object’s plug at the vehicle
end into test socket OUT2.
Attention!
!
Single-Phase Mains Connection
For correct phase connection, the
earthing contact plug must inserted
into the mains outlet such that only
the ICCB-IN1 L1 LED lights up.
The ICCB-IN1 PE LED lights up as well
in the case of polarity reversal.
➭
Connecting a 3-Phase, Mode 2/3 Charging Cable
➭
Insert the plug at the power supply end
of the test object into the outlet of the
adapter connected to IN3, or into IN2
(mode 3 charging cable).
➭
Insert the test object’s plug at the vehicle
end into test socket OUT2.
10 | 22
Fault Simulation
6
Fault Simulation
Mode 2 and 3 charging cables can be tested
with the test adapter with simulation of faults.
6.1.2 Simulated Interruption
➭
Start with the bottom rotary selec-
tor switch (S2) in the first ON position in
the orange area for single-phase inter-
ruption.
Switch through the positions, one after
the other, in the clockwise direction (ta-
ble from top to bottom):
Attention!
!
In any case, comply with the manu-
facturer’s recommendations con-
cerning measurements/tests to be
conducted in accordance with
DGUV, regulation 3.
➭
Rotary Test
switch Adapter
ICCB Action
The following faults can be simulated for
mains power supply to a mode 2 charging
cable:
LED PE
LED
L1
LED
•
•
Reversed wires
Interruption of individual conductors
(undervoltage detection)
PE to phase (interference voltage in the
protective conductor due to connection
of the phase conductor to the protective
conductor)
The test object’s
1
2
3
4
5
6
correct reaction to
the simulated fault
can be found in
the manufac-
•
turer’s operating
instructions.
The evaluation of the test object’s reaction to
each respective fault is strictly visual:
•
•
ICCB active or inactive
(indicator lamp on the ICCB)
Fault indication by means of LEDs on the
test adapter
6.1
Mode 2 Charging Cable
(Single-phase)
6.1.3 Simulation of PE to Phase (PE-U
)
EXT
➭
Switch mains power on by
➭
Set the bottom rotary selector switch
(S2) to the PE-U position.
turning the top rotary selector switch (S1)
to the Status C position.
EXT
Interference voltage is switched to PE with
touch protection.
The L1 LED and the PE LED light up.
6.1.1 Simulation of Reversed Wires
➭
Turn the bottom rotary selector
switch (S2) to the L1-PE position in
the orange area for single-phase wire re-
versal.
Rotary Test
switch Adapter
ICCB Action
L1
PE
LED
Instead of the L1 LED, the PE LED must light up.
LED LED
The test object’s
8
Rotary Test
ICCB Action
correct reaction to
the simulated fault
can be found in the
manufacturer’s op-
erating instruc-
tions.
switch Adapter
L1
PE
LED
LED LED
The test object’s
7
correct reaction to
the simulated fault
can be found in the
manufacturer’s op-
erating instructions.
11 | 22
Fault Simulation
6.2
➭
Mode 2 (3-phase) Charging Cable
Switch mains voltage on by turning the top rotary selector switch (S1) to the
Status C position (system state: vehicle ready for charging).
6.2.1 Simulated Interruption
➭
Start with the bottom rotary selector switch (S2) in the first ON position in the green area
for single-phase interruption.
➭
Switch through the positions, one after the other, in the clockwise direction (table from
top to bottom). Conductor interruption using a 3-phase, mode 3 charging cable as an
example:
Note
Tripping performance may deviate from this example in the case of ICCBs from other man-
ufacturers – adhere to the manufacturer’s test instructions!
Rotary Test Adapter
switch
ICCB Action
LED
L1
LED LED
N
PE
LED L2
L3
LED LED
The test object’s correct reaction to the simulated fault can
1
2
3
4
5
6
7
8
9
be found in the manufacturer’s operating instructions.
Regarding step 8: The N LED also lights up due to star con-
nection, although N is interrupted!
10
6.2.2 Simulation of PE to Phase (PE-U
)
EXT
➭
Set the bottom rotary selector switch (S2) to the PE-U position.
EXT
ICCB-IN1 Action
LED
Rotary switch Test Adapter
L1
LED LED
N
PE
LED L2
L3
LED LED
The test object’s correct reaction to the simu-
lated fault can be found in the manufacturer’s
operating instructions.
11
➭
12 | 22
Measurements/Tests with Test Instruments
PROFITEST MF XTRA, PROFITEST MF TECH,
PROFITEST MXTRA, PROFITEST MXTRA IQ,
PROFITEST MTECH+, PROFITEST MTECH+ IQ:
2-pole measuring adapter
PROFITEST PRIME, PROFITEST PRIME AC:
probes L(1) and PE(3)
SECUTEST ST PRO: 2-pole measurement
with test probes P1 and P2 (P1-P2)
Switch the test object’s ICCB on.
Perform the measurement as described
in the operating instructions for the re-
spective test instrument.
7
Measurements/Tests with Test
Instruments
7.1
General Information
When connecting a compatible test instru-
ment, additional measurements/tests are
possible which are described in this chapter.
Important information for the performance of
measurements/tests
➭
➭
Attention!
!
When turning the switch, pay atten-
tion to the exact snap-in position and
the corresponding optical signals of
the LEDs, as described in these oper-
ating instructions.
7.2.2 Mode 3 Charging Cable
➭
➭
➭
The rotary selector switch must be set to
the R position.
ISO
Connect the test object as described in
•
•
The test adapter must remain connected to
the mains for protective conductor measure-
Connect the test object to the ICCB-IN1
PE and ICCB-OUT1 PE sockets:
PROFITEST MF XTRA, PROFITEST MF TECH,
PROFITEST MXTRA, PROFITEST MXTRA IQ,
PROFITEST MTECH+, PROFITEST MTECH+ IQ:
2-pole measuring adapter
PROFITEST PRIME, PROFITEST PRIME AC:
probes L(1) and PE(3)
SECUTEST ST PRO: 2-pole measurement
with test probes P1 and P2 (P1-P2)
Perform the measurement as described
in the operating instructions for the re-
spective test instrument.
ments (
R
) at mode 2 charging cables by
PE
means of the respective test instrument.
The test adapter can remain connected
to the mains for protective conductor
measurements conducted at test
objects whose protective conductor is
not switched, as well as for insulation
measurements (Riso) with the respective
test instrument.
➭
7.2
Measurement of Protective Conduc-
tor Resistance (R )
LO
7.2.1 Mode 2 Charging Cable
7.3
Measuring Insulation Resistance (R ISO)
As opposed to the usual default setting for
low-resistance measurements, the device
under test does not have to be disconnected
from all sources of voltage in this case. Both
rotary selector switches (S1 and S2) have to be
set to the ON and the Status C positions, in
order to be able to activate the test object and
switch the protective conductor through. Prior
to a protective conductor resistance measure-
ment, an offset measurement must be carried
out (the offset value of the test adapter must
be taken into account during the test).
Measurement is carried out between ICCB-
OUT1-PE and OUT2-PE.
➭
➭
Attention!
!
Insulation resistance may only be
measured at voltage-free devices!
The upper-hand rotary selector
switch must be set to the mains dis-
connect position: R
INS
➭
Connect the test instrument to the ICCB-
IN1 socket for the mains-end of the cable
and the ICCB-OUT1 socket for the output
end of the cable of the device under test:
PROFITEST MF XTRA, PROFITEST MF TECH,
PROFITEST MXTRA, PROFITEST MXTRA IQ,
PROFITEST MTECH+, PROFITEST MTECH+ IQ:
2-pole measuring adapter
Connect the test object as described in
Connect the test object to the ICCB-IN1
PE and ICCB-OUT1 PE sockets:
PROFITEST PRIME, PROFITEST PRIME AC:
probes L(1) and PE(3)
13 | 22
Measurements/Tests with Test Instruments
SECUTEST ST PRO: 2-pole measurement
We recommend the following procedure in
order to prevent an upstream RCD (selective
RCD as well) from tripping, and to ensure
that actual tripping current or correct time to
trip is measured:
with test probes P1 and P2 (P1-P2)
➭
Connect the test object as described in
Mode 2 Charging Cable:
➭
Connect the respective test instrument
with pole (L1) to the voltage conducting
pole depending on the earthing contact
plug wiring of charging cable L1, N (ICCB-
OUT1), and with pole (PE) to N (ICCB-IN1).
(observe limit values and manufacturer’s
specifications):
➭
Connect the test instrument to the ICCB-
IN1/ICCB-OUT1 L1 or N and PE sockets,
one after the other:
PROFITEST MF XTRA, PROFITEST MF TECH,
PROFITEST MXTRA, PROFITEST MXTRA IQ,
PROFITEST MTECH+, PROFITEST MTECH+ IQ:
2-pole measuring adapter
PROFITEST PRIME, PROFITEST PRIME AC:
probes L(1) and PE(3)
SECUTEST ST PRO: 2-pole measurement
with test probes P1 and P2 (P1-P2)
Perform the measurement as described
in the operating instructions for the re-
spective test instrument.
PROFITEST MF XTRA, PROFITEST MF TECH,
PROFITEST MXTRA, PROFITEST MXTRA IQ,
PROFITEST MTECH+, PROFITEST MTECH+ IQ:
2-pole measuring adapter
PROFITEST PRIME, PROFITEST PRIME AC:
probes L(1) and PE(3)
➭
➭
Refer to the operating instructions for the
respective test instrument regarding per-
formance of the measurement.
Mode 3 Charging Cable
➭
Connect the respective test instrument
to the ICCB-IN1/ICCB-OUT1 L1, L2, L3, N
and PE sockets, one after the other:
PROFITEST MF XTRA, PROFITEST MF TECH,
PROFITEST MXTRA, PROFITEST MXTRA IQ,
PROFITEST MTECH+, PROFITEST MTECH+ IQ:
2-pole measuring adapter
PROFITEST PRIME, PROFITEST PRIME AC:
probes L(1) and PE(3)
SECUTEST ST PRO: 2-pole measurement
with test probes P1 and P2 (P1-P2)
7.4
Tripping Test with Nominal Residual
Current and Measurement of Time to
Trip at Mode 2 Charging Cable
(not possible with SECUTEST ST PRO)
Note
If you operate your test adapter at an
electrical system with a 30 mA
breaker, the mains RCD may be
tripped during the tripping test (trip-
ping current or time to trip) instead of
the charging cable’s ICCB. In this
case the ICCB-IN1 L1 LED goes out.
14 | 22
Protective Conductor Current Measurement (IPE) at Mode 2 Charging Cable
7.5
Guided Test Sequences
8
Protective Conductor Current
Measurement (IPE) at Mode 2
Charging Cable
(not possible with PROFITEST MXTRA
and PROFITEST MTECH+)
Test sequences are available for the test
instruments. The test instrument runs
through all test steps semi-automatically.
Each test step is evaluated and assessed by
the user (go/no-go) for later documentation.
It depends on the respective test instrument
whether or not and which test sequences
are available.
The test sequences are provided free of
charge for download from the myGMC por-
tal, however, it is the responsibility of the
user to import them into the test instrument.
Protective conductor current or bias current
may result in premature tripping of RCDs
used in ICCBs.
For this reason, the protective conductor is
led out of the housing as a loop between
surface mount sockets 4 and 6. This makes
it possible to measure any protective con-
ductor current with the help of a current
clamp transformer.
You therefore require an additional current
clamp transformer (e.g. METRACLIP 61).
➭
➭
Connect the test object as described in
Switch mains voltage on by turning the
top rotary selector switch (S1) to the Sta-
tus C position (system state: vehicle
ready for charging).
➭
Create a myGMC account and register
your test instrument (both free of
charge):
➭
➭
➭
Turn the bottom rotary selector switch
(S2) to the first ON position in the orange
area for single-phase interruption.
Close the jaws of the current clamp
transformer around the external protec-
➭
➭
Download the test sequence(s).
Transfer the test sequence(s) to your test
instrument. Complete information can be
found in the product documentation for
the respective test instrument.
Perform the test sequence with the test
instrument and the test adapter.
Detailed information on the test se-
quences can be found in the product
documentation of the respective test in-
strument.
tive conductor loop I .
PE
Read the measured value for protective
conductor current at the current clamp
transformer.
Protective conductor current should not
exceed 3.5 mA.
➭
15 | 22
Measuring RC (Testing of Resistance Coding) (not possible with SECUTEST ST PRO)
9
Measuring RC
10
Storage and Transport
(Testing of Resistance Coding)
(not possible with SECUTEST ST
PRO)
Attention!
Improper Storage
!
Damage to the product and measur-
ing error due to environmental influ-
ences
Store the instrument in a protected
location and only within the limits of
permissible ambient conditions. The
respective ambient conditions (tem-
perature, humidity etc.) can be found
Testing of specified resistance value R for
maximum current-carrying capacity of the
mode 2/3 charging cable per
DIN EN 61851-1 (VDE 0122-1)
Additionally, you require a standards-compli-
ant resistance measuring instrument.
C
➭
➭
Connect the test object as described in
Measure resistance between the PP and
PE sockets at ICCB-OUT1 with a stan-
dards-compliant measuring instrument.
Compare the measured resistance value
with the value indicated in the table
below (Table B3 off DIN EN 61851-1).
Note
If the test adapter has not been used
for a long period of time, the switches
may demonstrate increased contact
resistance depending upon storage
conditions.
If this is the case, actuate the
switches several times.
➭
Note
Observe the maximum current-carry-
ing capacity specified on the cable.
Vehicle state
Function Test Result
Attention!
!
Testing of resis- Resistance 13 A charging cable,
Improper Transport
tance coding for measure-
vehicle inlets and ment
1.5 k
20 A charging cable, 680
Damage to the product and measur-
ing error
Transport the instrument only within
the limits of permissible ambient con-
ditions (temperature, humidity etc.).
vehicle connec- with multi-
tors per
meter or
test instru-
ment
32 A charging cable, 220
IEC 61851
Table B.3
63 A charging cable, 100
16 | 22
Maintenance
11.2.1 Testing of Protective Conductor Resis-
tance R
11
Maintenance
PE
11.1 Cleaning of Housing
Test Adapter Contacting, Single-Phase (230 V)
Protective conductor resistance is tested
between the PE contact at the mains plug
(inlet plug N1) and PE contacts ICCB IN1
through IN4.
Attention!
Life endangering due to electric
shock!
!
The test adapter is operated with
electric power. Therefore there is a
general risk of electric shock which
can be fatal or cause severe injury.
– The test adapter must be voltage-
free both before and during cleaning.
Switch the test adapter off and dis-
connect it from the mains power
supply.
– The test adapter may not be con-
nected to a measuring circuit before
and during cleaning.
– Never immerse the test adapter in
water or other liquids.
Testing the Single-Phase Connection
➭
Turn the top rotary selector switch (S1) at
the test adapter to the Status C position
(system state: vehicle ready for
charging).
With the exclusion of the exceptions speci-
fied in the following table, a protective con-
ductor resistance R of less than 2 is
PE
permissible. This is due to the design of the
test adapter.
Rotary Test
switch Adapter
ICCB Protective
Conductor Re-
sistance R PE,
– Never touch the test adapter to
with wet or moist hands.
L1
PE
LED Permissible
LED LED
No special maintenance is required. Keep
outside surfaces clean and dry.
1
2
3
4
5
6
< 2 W
< 2 W
< 2 W
< 2 W
< 2 W
> 30 MW
Attention!
!
Unsuitable cleaning agents such as
aggressive or abrasive cleansers re-
sult in damage to the test adapter.
– Use a cloth for cleaning which has
been slightly dampened with water.
– Avoid the use of cleansers, abra-
sives or solvents.
11.2 Technical Safety Inspections
Testing per DGUV Regulation 3
Subject your test adapter to technical safety
inspections at regular intervals.
The test adapter is designed in accordance
with IEC 61010 as a protection category I
and II test instrument.
7
8
> 30 MW
> 30 MW
Testing of the protective conductor, insula-
tion resistance and touch current is
described in the following subsections.
Test Adapter Contacting, 3-Phase (400 V)
Protective conductor resistance is tested
between the PE contact at mains connec-
tion N2 and the PE contact at ICCB-IN1
(parallel to IN2, IN3 and IN4).
17 | 22
Maintenance
Testing the 3-Phase Connection
•
•
at the power supply end of the test
object, ICCB-IN1
at the output end of the test object, ICCB-
OUT1
➭
Turn the top rotary selector switch (S1) at
the test adapter to the Status C position
(system state: vehicle ready for
charging).
in each case against PE.
The usual limit values apply.
➭
Turn the bottom rotary selector switch
(S2) to each switch position*.
11.2.3 Touch Current Measurement
With the exclusion of the exceptions specified
in the following table, a protective conductor
Touch current measurement is conducted at
the screw connections of the IN sockets
(IN2, IN3 and IN4) with the standard limit val-
resistance R of less than 2
is permissible.
PE
This is due to the design of the test adapter.
ues (I < 0,5 mA).
B
Rotary
switch
Protective Conductor Resis-
tance R PE,
11.3 Fuse Replacement
The test adapter may only be operated with
fuses in flawless condition. Damaged or
defective fuses must be replaced immedi-
ately.
All fuses for neutral and phase conductors
are accessible from the outside.
Permissible
< 2 W
1
2
< 2 W
3
< 2 W
Attention!
Life endangering due to electric
shock!
!
4
< 2 W
The test adapter is operated with
electric power. Therefore there is a
general risk of electric shock which
can be fatal or cause severe injury.
– The test adapter must be voltage-
free both before and during changing
the fuse. Switch the test adapter off
and disconnect it from the mains
power supply.
5
< 2 W
6
< 2 W
7
< 2 W
8
< 2 W
– The test adapter may not be con-
nected to a measuring circuit before
and during cleaning.
9
< 2 W
10
11
> 30 MW
> 30 MW
Attention!
!
Only install fuses of a type that con-
forms to the details specified in the
No other fuses may not be used in
the test instrument.
The PE connection between OUT1 and
OUT2 must also be tested (< 2 ).
11.2.2 Testing Insulation Resistance
Testing is conducted in the respective switch
positions for Status B, C or E and in the
green ON position for 3-phase testing at
short-circuited L-N or L1-L2-L3-N contacts
(for earthing contact and CEE respectively)
•
at mains connections N1 and N2
18 | 22
Repair
F1, F2 and F3 (3-phase mains supply power sys-
tem N2)
Required materials:
12
Repair
If your instrument requires repair, please
contact our service department, see “Con-
–
A slotted screwdriver for opening the
fuse compartment
–
Replacement fuses in accordance with
the details specified in the Technical Data
Note
Loss of warranty and guarantee
claims!
Unauthorized modification of the tes-
ter is prohibited. This also includes
opening the meter.
If it can be ascertained that the tester
has been opened by unauthorized
personnel, no guarantee claims can
be honored by the manufacturer with
regard to personal safety, measuring
accuracy, compliance with applicable
safety measures or any consequen-
tial damages.
➭
Open the respective fuse compartment.
Insert the slotted screwdriver into the slot
on the fuse compartment and turn it
counterclockwise.
Pull out the fuse cover as soon as it pro-
trudes slightly from the housing.
Remove the defective fuse from the fuse
cover.
Insert the new fuse into the fuse cover.
Insert the fuse cover with the fuse back
into the fuse compartment. The two tabs
on the fuse cover must be inserted into
the recesses in the housing.
Insert the slotted screwdriver into the slot
on the fuse compartment and turn it
clockwise until the fuse compartment is
closed.
➭
➭
➭
➭
If the guarantee seal is damaged or
removed, all guarantee claims are
rendered null and void.
➭
The device may only be repaired or opened
by authorized, qualified personnel who are
familiar with the associated dangers.
Original replacement parts may only be
installed by authorized, qualified personnel.
The instrument may not be placed back into
operation until troubleshooting and repair
have been performed, and calibration and
dielectric strength have been tested and
approved at our factory or by an authorized
service center.
FLN (mains power supply system single-phase N1)
You require replacement fuses in accor-
dance with the details specified in the Tech-
➭
Open the fuse compartment. Press to-
gether the plastic wings of the fuse link
to this end and pull it up slightly.
As soon as the fuse link protrudes a little
bit, pull it out.
Remove the defective fuses from the
fuse link.
➭
➭
➭
➭
Insert the new fuses into the fuse link.
Insert the fuse link with the fuses into the
fuse compartment. The tab on the fuse
cover must be inserted into the recesses
in the housing.
➭
Press down the fuse link until it snaps
into place.
19 | 22
Contact, Support and Service
13
Contact, Support and Service
14
CE Declaration
Gossen Metrawatt GmbH can be reached
directly and simply – we have a single num-
The instrument fulfills all requirements of
applicable EU directives and national regula-
ber for everything! Whether you require sup- tions. We confirm this with the CE mark. The
port or training, or have an individual inquiry,
we can answer all of your questions here:
CE declaration can be found on our website:
+49-911-8602-0
Monday to
Thursday:
Friday:
8 a.m. to 4 p.m.
8 a.m. to 2 p.m.
Or contact us by e-mail at:
Do you prefer support by e-mail?
Measuring and Test Technology:
Industrial Measuring Technology:
Please contact GMC-I Service GmbH for
repairs, replacement parts and calibration*:
+49-911-817718-0
Beuthener Str. 41
90471 Nürnberg
Germany
*
DAkkS calibration laboratory per DIN EN ISO/IEC 17025
accredited by the Deutsche Akkreditierungsstelle GmbH under
reference number D-K-15080-01-01.
20 | 22
Disposal and Environmental Protection
limit value for the respective metal.
15
Disposal and Environmental
Protection
Please observe the owner’s or end user’s
responsibility with regard to deleting per-
sonal data, as well as any other sensitive
data, from old devices before disposal.
Old devices, electrical or electronic accesso-
ries and waste batteries (including recharge-
able batteries) used in Germany can be
returned free of charge to Gossen Metrawatt
GmbH or the service provider responsible
for their disposal in compliance with applica-
ble regulations, in particular laws concerning
packaging and hazardous goods. Waste
batteries must be handed over in discharged
state and/or with appropriate precautionary
measures against short-circuiting. Further
information regarding returns can be found
on our website.
Proper disposal makes an important contri-
bution to the protection of our environment
and the conservation of natural resources.
Attention!
!
Environmental Damage
Improper disposal results in environ-
mental damage.
Follow the instructions concerning
return and disposal included in this
section.
The following comments refer specifically to
the legal situation in the Federal Republic of
Germany. Owners or end users who are
subject to other regulations must comply
with the respective local requirements and
implement them correctly on site. Further
information can be obtained, for example,
from the responsible authorities or the local
distributor.
Packaging Materials
We recommend retaining the respective
packaging materials for the case that you
might require servicing or calibration in the
future.
Waste Electrical Equipment, Electrical or Elec-
tronic Accessories and Waste Batteries (including
rechargeable batteries)
Electrical equipment and batteries (including
rechargeable batteries) contain valuable raw
materials that can be recycled, as well as
hazardous substances which can cause
serious harm to human health and the envi-
ronment, and they must be recycled and
disposed of correctly.
Attention!
!
Danger of Asphyxiation Resulting
from Foils and Other Packaging Ma-
terials.
Children and other vulnerable per-
sons may suffocate if they wrap
themselves in packaging materials,
or their components or foils, or if they
pull them over their heads or swallow
them. Keep packaging materials, as
well as their components and foils,
out of the reach of babies, children
and other vulnerable persons.
The symbol at the left depicting a
crossed-out garbage can on
wheels refers to the legal obligation
of the owner or end user (German
electrical and electronic equipment act Elek-
troG and German battery act BattG) not to
dispose of used electrical equipment and
batteries with unsorted municipal waste
(“household trash”). Waste batteries must be
removed from the old device (where possi-
ble) without destroying them and the old
device and the waste batteries must be dis-
posed of separately. The battery type and its
chemical composition are indicated on the
battery’s labelling. If the abbreviations “Pb”
for lead, “Cd” for cadmium or “Hg” for mer-
cury are included, the battery exceeds the
In accordance with German packaging law
(VerpackG), the user is obligated to correctly
dispose of packaging and its components
separately, and not together with unsorted
municipal waste (“household trash”).
Private end consumers can dispose of pack-
aging free of charge at the responsible col-
lection point. Packaging which is not subject
to so-called system participation is returned
to the appointed service provider. Further
information regarding returns can be found
on our website.
21 | 22
©Gossen Metrawatt GmbH
Prepared in Germany • Subject to change / errors excepted • A pdf version is available on the Internet
All trademarks, registered trademarks, logos, product names and company names are the property of their respective
owners.
Gossen Metrawatt GmbH
Südwestpark 15
90449 Nürnberg
Germany
Phone +49 911 8602-0
Fax +49 911 8602-669
E-Mail [email protected]
|