Introduction  
The GE Protimeter Aquant BLD5760 moisture detector is used to  
assess and monitor the relative moisture level of non-conductive  
and porous materials: such as plaster, masonry, concrete and  
fiberglass (GRP). It is an ideal instrument for making rapid surveys  
of solid walls and floors and to pinpoint areas of concern in tiled  
shower cubicles, GRP boat hulls, motor homes and caravans etc.  
The Aquant BLD5760 has a digital display that is synchronized  
with a color coded LED scale. The digital display shows the relative  
moisture level of the material under investigation on a scale of 0-  
1000 and the LED scale indicates moisture condition of the material.  
The color-coded zones signify the moisture condition of the material under investigation. The green,  
yellow, and red color-coded zones indicate the following:  
Material in the green zone is in a safe air-dry condition.  
In the yellow zone moisture levels are higher than normal but not critical; further investigation is  
recommended.  
The red zone represents excessive moisture levels. If sustained, red zone moisture levels will lead  
to decay in organic materials.  
1 Operation  
The Aquant BLD5760 moisture detector has a radio frequency transceiver located in the bulge on it’s  
underside to give relative readings within the material against which it is held up to a nominal depth of  
3/4 in. (19 mm). The depth of measurement is subject to the characteristics of the material under test.  
To use the Aquant, do the following:  
1. Hold the Aquant away from surfaces as shown and switch ON by pressing  
momentarily. The instrument may now be used to take measurements  
or zeroed as appropriate.  
2. Zero the instrument (with respect to the environment in which it is about to  
be used) by pressing and holding for 3 seconds, until the word nuL”  
appears in the display.  
3. Release ; nuLwill flash for a few seconds and then disappear from  
the screen. The Aquant is now zeroed.  
INS5760, Rev. A1  
April 2007  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Scroll across the display and record the following information:  
a. part number (“bLd5760”)  
b. firmware date in the form yy-mm-dd (e.g. “04-05-20”)  
c. calibration code (e.g. E1-1).  
Note: Once the scrolling is complete, display shows “0 = 0” meaning that option 0 is set to 0. The  
and buttons may now be used to change the instrument options and settings,  
respectively, as detailed in Table 1 below.  
TABLE 1. User Set-Up Options  
When  
Option No. Is:  
And Setting  
Number Is:  
Aquant Set-Up Is:  
0
0
1
Instruments settings are not changed from previous settings  
Instrument default settings loaded - switches OFF  
automatically after 1 minute, beeper is activated, factory-set  
zeroing  
1
0
1
2
Beeper is switched OFF  
Beeper beeps when switching from one mode to the other  
Beeper beeps with increasing frequency from nominal value  
of 170  
2
0
Auto switch OFF is not active. Instrument can only be  
switched OFF by pressing  
and holding it for over 3 sec.  
1
2
3
Auto switch OFF is active. Instrument switches OFF after 1  
minutes  
Auto switch OFF is active. Instrument switches OFF after 2  
minutes  
Auto switch OFF is active. Instrument switches OFF after 3  
minutes  
Note: The instrument options can only be changed in numerical order (0, 1, and 2 respectively).  
Save setting changes and exit the Set-up mode by pressing  
following example:  
again, as in Step 4 of the  
Example:  
To switch the audible beeper OFF and set the auto switch OFF time to 3 minutes, complete the  
following steps:  
1. Press and hold followed by  
to enter Set-up mode.  
2. When display shows 0 = 0, press  
once to select beeper options (1), then press until the  
display shows 1 = 0.  
GE Protimeter Aquant  
Page 3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Press  
once again to select auto switch off options (2), then press until the display shows  
2 = 3.  
4. Press  
to save settings, exit Set-up mode and return to operational modes.  
6 Error Messages  
When zeroing if the error message E32 appears on the instrument display, signifying that the  
instrument reading is beyond the zeroing range of 130.  
7 Care and Maintenance  
When the Aquant is not being used, then do the following:  
Store the Aquant in the carry case (provided with the kit) in a stable and dry environment.  
Remove the battery, if the Aquant will not be used for extended period of time.  
Replace the battery when the low battery symbol appears on the display.  
The information contained in this leaflet is given in good faith. As the method of use of the  
instrument (and its accessories) and the interpretation of the readings are beyond the control of  
the manufacturers, they cannot accept responsibility for any loss, consequential or otherwise,  
resulting from its use.  
USA  
1100 Technology Park Drive  
Billerica, MA 01821-4111  
Ireland  
Sensing House  
Shannon Free Zone East,  
Shannon, County Clare  
®
Protimeter Aquant is a registered trademark of GE  
Page 4  
GE Protimeter Aquant  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Sony Camera Lens VCLECF1 User Manual
Sony Headphones DR BT101IK User Manual
Sony Robotics ERF 210AW02 User Manual
Speedotron Power Supply D1204 User Manual
Strong Pools and Spas Hot Tub The Antigua User Manual
Sunbeam Blood Pressure Monitor 7655 10 User Manual
SVA Flat Panel Television ED4228P User Manual
Swann TV Receiver SW112 DVB User Manual
TANDBERG Network Card D1417201 User Manual
Tannoy Cell Phone K770I User Manual