Sunbeam Blood Pressure Monitor 7655 10 User Manual

Home  
Finger Blood Pressure Monitor  
Instruction Booklet  
Read All Instructions Carefully Before Use  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Save These Instructions  
DO’S AND DON’TS  
FOR ACCURATE BLOOD PRESSURE READINGS  
1. DO Read instructions carefully before using Monitor.  
2. DO Rest arm on a flat surface so that finger is at heart level  
when taking measurements.  
3. DO Remain calm, quiet and still when measurements are being  
taken.  
4. DO Make sure that the Finger Cuff is adjusted properly according  
to your instruction manual.  
5. DO Insure that the exhaust velocity is 2-5 mmHg per second at  
the systolic point each time a measurement is taken.  
6. DO Wait at least 10 minutes between taking measurements.  
7. DO Pump unit 30 to 60 mmHg above normal systolic before beginning measurement.  
8. DO Consult your physician before adjusting your medication based on readings from  
this Monitor.  
9. DO Use fresh alkaline batteries.  
IO. DO Use a dry, soft cloth to clean the instrument.  
11. DO use your left index finger when taking measurement. Use of an alternate finger may  
result in an inaccurate reading.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
12. DO make sure that your finger is warm.  
1. DO NOT move arms or hands when measurement is being taken.  
2. DO NOT talk or chew gum when measurement is being taken.  
3. DO NOT take a series of measurements without waiting at least 10 minutes  
between  
measurements.  
4. DO NOT adjust medication based on readings from this Monitor.  
5. DO NOT make any adjustments to unit; there are no user serviceable parts. Return to  
Service Station for repair.  
6. DO NOT smoke during or at least 30 minutes prior to taking blood pressure measurements.  
7. DO NOT drop or otherwise mechanically shock the Monitor. This is a delicate  
precision device.  
8. DO NOT use thinner, alcohol, benzene or wet cloth to clean unit.  
9. DO NOT subject to extremes in temperature, humidity, direct sunlight or dust.  
Manufactured to meet specifications set by the Association for the Advancement of  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Medical Instrumentation (AAMI)  
WHAT IS BLOOD PRESSURE?  
Measurement of the rate of pressure exerted by the blood upon the  
walls of the blood vessels. The heart contracts and then  
relaxes, producing the two levels of blood pressure.  
Bloodpressure:  
The highest point of blood pressure; level of blood pressure  
produced as your heart beats.  
Systolic:  
Diastolic:  
The lowest point of blood pressure; level of blood pressure  
produced as your heart rests between beats.  
Millimeters of Mercury (mmHg):  
Blood pressure measurements are measured in terms of the  
height of mercury in a column. Measurements are written one  
over the other. For example, a systolic pressure of 120 mmHg and  
diastolic pressure of 80 mmHg is expressed as “120 over 80.”  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Blood pressure varies constantly and may change due to:  
l
l
l
-----  
l
l
physical trauma  
mental anxieties  
smoking  
exercise  
eating  
drinking caffeinated or alcoholic beverages  
Only a qualified physician can determine whether your blood pressure is normal.  
is high blood pressure, and is most common among adults and the  
elderly; left unattended, it can cause serious health problems such as a  
stroke, heart attack, etc. Therefore, it is important to monitor blood  
pressure on a continued and timely basis.  
Hypertension:  
Only a physician is qualified to interpret your blood pressure  
measurements, and no device can replace regular medical exami-  
nations by your physician. It is recommended that your physician  
review your procedure for using this blood pressure monitor. Your  
physician should verify blood pressure measurements before  
DomwnalokadinfrgomaWdwjwu.sSotmmaenunatlss.cotmo. Amll eMdaniucaalstSioeanrc.h And Download.  
IMPORTANT:  
BLOOD PRESSURE RANGES  
In young, healthy adults the normal blood pressure range is between 100 and 140 mmHg (milli-  
meters of mercury) for SYSTOLIC pressure, and between 60 and 90 mmHg for DIASTOLIC  
pressure.  
The range of pressure can vary considerably depending on age and physical condition.  
ONLY A QUALIFIED PHYSICIAN CAN DETERMINE WHAT IS “NORMAL” FOR A  
PARTICULAR  
PERSON.  
activity  
Your reading can also vary during the .day and depending on your physical condition and  
level. In general, readings will be lower when you are calm and rested, and higher following  
physical activity. There will be some variance in consecutive readings or readings from one arm  
to the other.  
For these reasons, it is recommended that blood pressure be measured after sitting or  
resting quietly for 30 minutes or more following any strenuous activity.  
To minimize variations, readings should be taken at similar times and under similar conditions of  
stress. Record daily readings for accurate comparisons.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PARTS OF YOUR BLOOD PRESSURE MONITOR  
Cuff Size  
Indicator  
Finger Cuff  
\
Start  
Button  
On/Off  
Button  
Blood Pressure and  
Pulse Rate Display  
Download from Www.Somanuals.com. All Manuals Search And Download.  
DESCRIPTION OF DISPLAY MARKS  
CORRECTIVE ACTION  
CONDITION/CAUSE  
,
DISPLAY MARK  
Measurement in progress; re-  
main quiet.  
Heart mark appears during  
measurement conditions and  
flashes when pulse is de-  
tected.  
Measurement in progress  
Automatic exhaust occurs.  
Mark flashes when power is  
turned on and there is air re-  
maining in the Cuff. Mark  
flashes at the completion of  
the measurement until ex-  
haust is complete.  
Exhausting  
Set the ON/OFF Button to Off.  
Set it back to ON, and then  
start the measurement again  
when “0” appears.  
Measurement began, but the  
pressure was insufficient.  
Insufficient Pressure  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appears when the battery is  
excessively low.  
Replace all four batteries with  
new ones. (4  
-
AAA  
Replace Batteries  
Alkaline batteries)  
Appears when the blood  
pressure cannot be mea-  
sured accurately.  
Press ON/OFF Button to turn  
OFF. Press again to ON and  
start the measurement again  
when “0” appears.  
Measurement  
Error  
The pulse is very weak; or  
the finger is cold; or Finger  
Cuff is too loose.  
If you have poor circulation in  
your hand, you will not get a  
reading. Warm up your hand  
and retry measurement.  
Tighten Finger Cuff and retry  
measurement.  
E
r
r
Download from Www.Somanuals.com. All Manuals Search And Download.  
BATTERY INSTALLATION  
4 AAA alkaline batteries are required. Remove the  
battery cover and insert batteries into the battery  
. compartment on back of monitor. (Fig.1) Be sure that  
you insert the batteries according to the polarities  
specified on the bottom of the battery compartment. If  
Monitor is to be stored for extended periods, remove the  
batteries.  
Figure 1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
HOW TO USE YOUR SUNBEAM DIGITAL  
BLOOD PRESSURE MONITOR  
1. Slide the Cuff Size Adjuster to the far left (L position)  
to open the finger cuff fully.  
2. Insert your left index finger fully into the Finger Cuff.  
(Use of an alternate finger will result in an improper  
reading) (Fig.2) Be sure your finger is warm and  
relaxed. (If your finger is cold an improper reading may  
result)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure  
2
3. Press the ON/OFF Button. Three sets of symbols will  
appear in rapid sequence on the display. (Fig. 3) Wait  
for the "O" symbol to appear in the display before  
starting measurement, indicating the unit is ready to  
measure.  
4. Slide the Cuff Size Adjuster to the right until the Cuff  
Size Indicator light turns on. When the red light turns  
on, this means that the Finger Cuff is too tight. If your  
fingers are slender, the smallest finger cuff position may  
be the most suitable. In this case, the red Too Tight  
Indicator will not light up. An inaccurate blood pressure  
reading may result from improper Finger Cuff adjust-  
ment.  
5. Hold the Monitor steady at heart level with one or  
both hands. Be sure that when holding the Monitor, your  
left hand is relaxed. Press the Start Button. The Finger  
Cuff will automatically inflate to a pressure of about 200  
mmHg. If a higher starting point than 2OOmmHg is  
required, the monitor will stop inflation automatically,  
then resume inflation after determining the correct  
Figure 3  
starting point.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Remain still after inflation is completed. The Monitor  
will then begin deflation of the Finger Bladder automati-  
cally. Pressure readings will be indicated in the display  
as the pressure in the Finger Bladder decreases. As the  
Monitor continues to deflate the Finger Bladder,  
a
heart  
will start to flash in the display. (Fig. 4)  
7. The Monitor will deflate automatically when measure-  
ment is complete, and the resulting blood pressure  
readings will be indicated in the display. Blood pressure  
readings (Systolic on the left, Diastolic on the right), will  
be displayed alternately with the pulse (Pul) reading.  
(Fig. 5)  
Figure 4  
8. Display of measurement results will continue until the  
power is turned off. The Monitor will shut off automati-  
cally after approximately  
2
minutes.  
9. Wait at least 10 minutes between readings. To  
discontinue measurement for any reason, press the  
a
ON/OFF Button. This causes the finger cuff to deflate  
rapidly. Before removing your finger from the Finger  
Cuff, slide the Cuff Adjuster to the left. Failure to do  
Figure 5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
so could result in damage to the Finger Cuff.  
SPECIFICATIONS  
Model: 7655-10  
Power Source: Type AAA batteries (4)  
Battery Life: Approx. 6 months with 3 min. usage per day  
Power Consumption: 600mW  
Type: Oscillometric  
Measurement Range:  
0
-
300mmHg (Pressure)  
Operating Environment:  
-
-
40  
-
150 pulses/minute (Pulse)  
30  
Storage Environment: -20°C  
30  
-
85%RH  
Accuracy: +I-3mmHg or +/-2% of reading whichever  
-
is greater  
-
95% RH  
x
Dimensions: 4”(W) x  
Pulse:  
+/- 5% of reading  
Finger Circumference Range: Approx. 2-l/8” to 3”  
Weight w/batteries: Approx. 240gm.  
Origin: Made in Japan  
Inflation: Automatic, using an electronic pump  
Deflation: Automatic pressure release valve  
Download from Www.Somanuals.com. All Manuals Search And Download.  
If you have any questions about your new blood pressure monitor, please call our toll free number for  
assistance at l-800-621 -8854.  
LIMITED ONE YEAR WARRANTY  
This product is warranted for one year from the date of purchase against defects in material and workmanship. If the product  
fails to operate during this period, return it, prepaid, to Service Center 117 Central Industrial Row Purvis, MS 39475  
for repair or replacement without charge at the manufacturer’s option. This warranty gives you specific legal  
rights. and you may also have other rights which vary from state to state. Warrantor: Sunbeam Home Comfort  
@Sunbeam TM Health at Home, 01992 Sunbeam Home Comfort Distributed by Sunbeam Home Comfort, Schaumburg, IL 60173  
P.N. 47834-025-000  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Sharp Calculator EL T100W User Manual
Sharp PDAs Smartphones YO 190 User Manual
Shure Automobile Parts a15HP User Manual
Signature Housewares Washer SIG100 97 User Manual
Socket Mobile Network Card E300 User Manual
Sony Car Speaker SS SP40FW User Manual
Sony Flat Panel Television XBR65X900C User Manual
Sony Speaker System SA VS310 User Manual
Sterling Model Vehicle 6828810700 User Manual
Sunbeam Slow Cooker SCRI500 I User Manual