Omega Binding Machine 3200 User Manual

FMA 3200/3200ST/3400/3400ST Series  
Thermal Mass Flow Controllers  
Download from Www.Somanuals.com. All Manuals Search And Download.  
READ THIS MANUAL COMPLETELY BEFORE ATTEMPTING TO CONNECT OR  
OPERATE YOUR FLOW SENSOR. FAILURE TO DO SO MAY RESULT IN INJURY  
TO YOU OR DAMAGE TO THE FLOW CONTROLLER.  
T A B L E  
O F C O N T E N T S  
A. Introduction .............................................................................................4  
1. Unpacking...........................................................................................4  
2. Product Overview And Principle Of Operation..........................................4  
B. Installation...............................................................................................5  
1. General Considerations.........................................................................5  
2. Mounting The flow controller ...............................................................7  
3. Tubing Connections..............................................................................7  
5. Electrical Connections ...........................................................................8  
a) Overview  
....................................................................................8  
b) Connecting The 6 Pin Mini Din Connector............................................9  
c) Connecting The 6 Pin Mini Din Connector & FMA 3000C Cable ...........10  
d) Connections For The 9 Pin D Sub Connector ......................................11  
e) Connections For The 15 Pin D Sub Connector.....................................12  
f) Using a 0-5VDC Output / Input Power Adapter Package.......................13  
C. Operation .............................................................................................14  
1. Warm-Up ..........................................................................................14  
2. Verification Of Zero ............................................................................14  
3. Flow Readings....................................................................................14  
4. Changing The Flow Rate Set-Point (Using An External Voltage Source) .......15  
5. Changing The Flow Rate Set-Point – FMA3400 / 3400ST Only.................16  
6. Power Save Mode...............................................................................16  
7. Zero Adjustments................................................................................17  
8. Recalibration......................................................................................17  
9. Changing The Calibration Gas – FMA3400 / 3400ST Only.....................17  
D. Maintenance And Product Care................................................................18  
1. General.............................................................................................18  
2. Returning Units For Repair Or Recalibration............................................18  
E. Specifications.........................................................................................19  
F. Dimensions............................................................................................20  
G. Gas K Factors........................................................................................22  
H. Trouble Shooting Guide ..........................................................................23  
M-4271/0707, pg. 3 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
A.  
Introduction  
1. Unpacking  
All units are suitably packaged to prevent damage during shipping. If  
external damage is noted upon receipt of the package, please contact  
Omega Engineering immediately.  
Open the package from the top, taking care not to cut too deeply into the  
package. Remove all the documentation and contents. Take care to  
remove all the items and check them against the packing slip. The  
products should also be checked for any concealed shipping damage. If  
any shortages or damage is noted, please contact Omega Engineering to  
resolve the problem.  
Typical Contents of Box: Controller, Calibration Certificate & Manual  
FMA 3200/3200ST shown; FMA 3400/3400ST have an integrated display.  
Caution: Take care not to drop your controller. Read the  
installation section of this manual before providing power or  
tubing connections to the unit. Any damage caused by improper  
installation or careless handling will not be repaired under  
warranty (see limited warranty on page 25 for more details).  
2. Product Overview and Principle of Operation  
The FMA 3200/3400 Series Mass Flow Controllers from Omega  
Engineering are capable of measuring and controlling the flow of virtually  
any clean, dry gas as low as 0-20 sccm or as high as 0-10 l/min.  
Repeatable results are achieved using a patented thermal mass flow  
M-4271/0707, pg. 4 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
sensor design. This proven design minimizes zero drift while maintaining  
fast response and linear outputs with virtually no maintenance.  
The FMA 3200/3400 Series  
utilizes thermal flow sensing  
technology. A portion of the  
gas flowing through the unit is  
redirected into a small sensor  
tube. This tube has two coils  
on the outside. The first coil  
introduces a small amount of  
heat into the gas stream. As  
the gas passes through the  
tube heat is transferred from  
one coil to the other. The flow  
rate is proportional to the  
amount of heat transfer. Smart  
electronics analyze the amount  
of temperature change in the  
second coil and provide a  
linearized analog output. A patented system insures that the zero remains  
stable and the sensor is extremely repeatable.  
Flow in the FMA 3200/3400 Series is controlled by a proportional solenoid  
valve with active servo electronics. The flow measurement signal is  
analyzed by micro-processor controlled electronics and compared to a set-  
point. Adjustments are then made to the valve in order to achieve the  
required flow rate. The set point can be either externally input via a 0-  
5VDC signal or in the case of the FMA 3400/3400ST Series it can be input  
manually on the unit.  
The output of the thermal mass flow sensor is directly related to the  
specific heat characteristic of the gas being measured. A sensor is  
calibrated for one gas but may be used with other gases by applying a  
correction factor to the output. The calibration gas for each specific flow  
controller is detailed on the product label.  
B.  
Installation  
Caution: Do not exceed the pressure, temperature or power  
operating ranges detailed in the SPECIFICATIONS section of this  
manual. Omega Engineering shall not be liable for any damage  
or injury caused by incorrect operation of their products.  
1. General Considerations  
It is recommended that a safety shut-off valve be installed upstream  
(before) of the controller.  
All wetted parts should be checked for compatibility with the gas to be  
used. If there are any incompatibilities eg. highly corrosive gas, then the  
M-4271/0707, pg. 5 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
unit may be damaged or fail prematurely. Such damage will not be  
repaired under warranty.  
Units should be installed in a clean, dry environment with an ambient  
temperature that is as stable as possible. Avoid areas with strong magnetic  
fields, strong air flows or excessive vibration.  
In order to operate the differential pressure across the controller should be  
in the range 15-45psid (1-3 bar). For optimum performance a differential  
pressure of 25psid is recommended.  
For Example, consider the following system:  
The differential pressure across the flow controller in this system would be  
100 psi – 14 psi = 86 psid. Consequently the flow controller would NOT be  
able to control flow. For the unit to operate at optimum performance the  
supply pressure from the gas cylinder would need to be lowered to 39 psig  
to give 39 psi -14 psi = 25 psid.  
M-4271/0707, pg. 6 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Mounting the Flow Controller.  
The FMA 3200/3400 Series controllers have no particular orientation or  
installation requirements so may be mounted in any convenient position.  
It is recommended that units be fixed to a suitable substrate using the two  
4-40 mounting holes provided.  
Mounting View from Bottom  
(mounting hardware not included with sensor)  
3. Tubing Connections  
All tubing must be clean, dry and purged with clean dry air before  
installation of the FLO-CONTROLLER®.  
If the gas to be used may contain particles then a filter (20 microns or  
less) should be installed upstream of (before) the unit.  
When connecting the sensor to the tubing, take care not to over-tighten  
the fittings or leaking may occur.  
Caution: Only use the fittings factory installed on the unit. If  
the fittings are removed the calibration of the unit may be  
effected and leaking may occur. If different fittings are required  
please contact the Omega Engineering Customer Service  
Department for assistance.  
M-4271/0707, pg. 7 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Electrical Connections  
Caution: Incorrect wiring may cause severe damage to the unit.  
Applying an AC voltage (115VAC or 230VAC) directly to the unit  
will cause damage. Read the following instructions carefully  
before making any connections.  
a) Overview  
The FMA 3200/3400 Series provides a 0-5VDC analog output proportional  
to the flow rate. This output may be connected to a display, data  
acquisition system or voltmeter with an impedance of greater than 2.5 k  
(kilo ohms).  
The flow controller needs to be supplied with a 0-5 VDC set point signal to  
enable control. On the FMA 3400/3400ST Series this may be generated  
internally by altering the set-point potentiometer on the front panel of the  
unit.  
A stable D.C. power supply is required to operate the unit. The voltage  
and current requirements depend on the configuration of the unit. Full  
details may be found in the Specification section of this manual.  
Connecting wires should be as short as possible to avoid voltage drops.  
Twisted conductor cable should be used if the length of the wiring is to be  
longer than 2 meters.  
Units are supplied with either a 6 pin mini DIN type connector (requires  
mating cable assembly), a 9 Pin D Sub connector or 15 Pin D Sub  
connector.  
Caution: Cutting off the integrated connectors on the unit IS  
NOT RECOMMENDED and will void the product warranty. Mating  
cables should be ordered along with each unit.  
Electrical connections to the units are made as detailed in the following  
sections.  
M-4271/0707, pg. 8 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
b) Connecting The 6 Pin Mini Din Connector  
Using a suitable mating connector the pins of the integrated connector  
should be wired as follows:  
Connecting To The Integrated 6 Pin Connector  
Pin Out of Integrated  
Connector  
Pin 2 should be connected to the Positive of the power source.  
Pin 6 should be connected to the Negative (Ground) of the power source.  
Pin 3 provides the signal output and should be connected to the positive  
terminal of the display, data acquisition system or voltmeter.  
Pin 1 is the signal negative (ground) and should be connected to the  
negative (Ground) terminal of the display, data acquisition system or  
voltmeter.  
Pin 4 provides the input signal and should be connected to the positive  
terminal of the voltage source. The (0-5VDC) voltage control signal should  
be supplied from a low impedance source.  
Pin 5 is the input signal negative (ground) and should be connected to  
the negative (Ground) terminal of the voltage source.  
Caution: Avoid high voltage static discharges to the input signal  
connection. Do not short the input/output signal wires or allow  
them to contact the power wires at any time. DAMAGE WILL  
RESULT!  
M-4271/0707, pg. 9 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
c) Connecting The 6 Pin Mini Din Connector & FMA 3000C Cable  
The two mating connectors should be pushed together and the pigtail  
leads wired as follows:  
Connecting To The Integrated 6 Pin Connector Using A FMA 3000C Cable  
The RED wire should be connected to the Positive of the power source.  
The BLACK wire should be connected to the Negative (Ground) of the  
power source.  
The ORANGE wire provides the signal output and should be connected to  
the positive terminal of the display, data acquisition system or voltmeter.  
The BROWN wire is the signal negative (ground) and should be  
connected to the negative (Ground) terminal of the display, data  
acquisition system or voltmeter.  
The YELLOW wire provides the input signal and should be connected to  
the positive terminal of the voltage source. The (0-5VDC) voltage control  
signal should be supplied from a low impedance source.  
The GREEN wire is the input signal negative (ground) and should be  
connected to the negative (Ground) terminal of the voltage source.  
The wire colors above describe the pigtail leads of the FMA 3000C cable  
assembly and may not correspond with the internal wiring of your flow  
sensor.  
Caution: Avoid high voltage static discharges to the input signal  
connection. Do not short the input/output signal wires or allow  
them to contact the power wires at any time. DAMAGE WILL  
RESULT!  
M-4271/0707, pg. 10 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
d) Connections For The 9 Pin D Sub Connector  
Using a suitable mating connector the pins of the integrated connector  
should be wired as follows:  
Connecting To The Integrated 9 Pin Connector  
Pin Out of Integrated  
Connector  
PIN 3 should be connected to the Positive of the power source.  
PIN 4 should be connected to the Negative ( Ground ) of the power  
source.  
PIN 2 provides the signal output and should be connected to the positive  
terminal of the display, data acquisition system or voltmeter.  
PIN 8 is the signal negative (ground) and should be connected to the  
negative (Ground) terminal of the display, data acquisition system or  
voltmeter.  
Pin 6 is the input signal and should be connected to the positive terminal  
of the voltage source. The (0-5VDC) voltage control signal should be  
supplied from a low impedance source.  
Pin 7 is the input signal negative (ground) and should be connected to  
the negative (Ground) terminal of the voltage source.  
Pins 1, 5, and 9 are not used.  
Caution: Avoid high voltage static discharges to the input signal  
connection. Do not short the input/output signal pins or allow  
them to contact the power connections at any time. DAMAGE  
WILL RESULT!  
M-4271/0707, pg. 11 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
e) Connections For The 15 D Sub Connector  
Using a suitable mating connector the pins of the integrated connector  
should be wired as follows:  
Connecting To The Integrated 15 Pin Connector  
Pin Out of Integrated  
Connector  
PIN 7 should be connected to the Positive of the power source.  
PIN 5 should be connected to the Negative ( Ground ) of the power  
source.  
PIN 2 provides the signal output and should be connected to the positive  
terminal of the display, data acquisition system or voltmeter.  
PIN 10 is the signal negative (ground) and should be connected to the  
negative (Ground) terminal of the display, data acquisition system or  
voltmeter.  
Pin 8 is the input signal and should be connected to the positive terminal  
of the voltage source. The (0-5VDC) voltage control signal should be  
supplied from a low impedance source.  
Pin 1 is the input signal negative (ground) and should be connected to  
the negative (Ground) terminal of the voltage source.  
Pins 3, 4, 6, 9, 11, 12, 13, 14 and 15 are not used.  
Caution: Avoid high voltage static discharges to the input signal  
connection. Do not short the output signal pins or allow them to  
contact the power connections at any time. DAMAGE WILL  
RESULT!  
M-4271/0707, pg. 12 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
f) Using a 0-5VDC Input / Output Power Adapter Package.  
An optional 0-5VDC Input / Output Power Adapter Package is available for  
use with the FMA 3200/3400 Series. This consists of a power source  
(115VAC or 230VAC), a connection hub and two cable assemblies with pig-  
tail (soldered wire) ends. This should be assembled as shown in the  
following diagram.  
Assembling a FMA 3215PW Power Adapter Package  
(the FMA 3223PW Power Adapter Package is similar)  
The RED connector should be inserted in the RED socket on the  
connection hub. The WHITE connector should be inserted in the WHITE  
socket on the connection hub.  
The cable with a RED connector provides the input signal. The RED wire  
of this cable should be connected to the positive terminal of the voltage  
source. The (0-5VDC) voltage control signal should be supplied from a low  
impedance source. The bare wire of this cable assembly is the input  
signal negative (ground) and should be connected to the negative  
(Ground) terminal of the voltage source.  
The cable with a WHITE connector provides the signal output. The  
WHITE wire should be connected to the positive terminal of the display,  
data acquisition system or voltmeter with an impedance of greater than  
2.5 k(kilo ohms). The bare wire of this cable assembly is the signal  
negative (ground) and should be connected to the negative (Ground)  
terminal of the display, data acquisition system or voltmeter.  
Caution: Avoid high voltage static discharges to the input signal  
connection. Do not short the output signal wires or allow them to  
contact the power wires at any time. DAMAGE WILL RESULT!  
M-4271/0707, pg. 13 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
C.  
Operation  
1. Warm Up  
Before applying power to the unit check all tubing and electrical  
connections. Once correct installation is verified switch on the power. The  
unit should then be allowed to warm up for 5 minutes before gas pressure  
is applied.  
2. Verification of Zero  
Flow through the unit should be stopped by sealing or capping the inlet of  
the controller. It is not adequate to only stop flow by turning off the gas  
supply or closing a valve as there may be a leak in the system. This would  
give a false reading.  
After 5 minutes, the zero should be stable when there is no flow through  
the unit. If after 10-15 minutes the output is still not zero volts (within  
±0.05 volts) the unit should be adjusted as detailed in section C part 6.  
It should be noted that power supply voltage variations and changes in  
ambient temperature can have an effect on zero readings.  
3. Flow Readings  
Each controller is factory calibrated for a specific flow range and gas (or  
gas mixture). The calibration gas and flow range are shown on the unit’s  
label and calibration certificate.  
By monitoring the voltage output signal it is possible to determine the flow  
rate of the gas. Units are configured so that an output signal of 5.0VDC is  
provided when the maximum flow (i.e. Full Scale flow) is passing through  
the unit. The output signal is linear and scaleable enabling calculation of  
flow rates with in the sensor’s range. For example:  
For a flow range of 0-500sccm:  
At 500sccm the output signal would be 5VDC  
If the output signal were 3.5VDC then the flow rate would be:  
500 ÷ 5 × 3.5 = 350sccm  
If the maximum flow rate is exceeded non-linear and inaccurate readings  
will result.  
Units may be used for gases other than the calibration gas. In this case a  
“K Factor” would need to be applied and a corrected value calculated using  
the following formula:  
M-4271/0707, pg. 14 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Q1 / Q2 = K1 / K2  
Q1 is the flow rate of the new gas  
Q2 is the flow rate of the original calibration gas  
K1 is the K factor of the new gas  
K2 is the K factor of the original calibration gas  
Q1 = (K1 / K2) Q2  
If K2 is larger than K1 then linear results will only be achieved if the unit  
does not exceed 5(K1/ K2)VDC for the full scale output.  
Example 1  
For a 0-200sccm unit calibrated for air the flow at 5.0VDC would be  
200sccm. The K factor for air is 1. If the unit is used with Helium (K factor  
1.454 relative to air) then the flow at 5VDC (i.e. the maximum flow) would  
be (1.454/1)200 = 290.8 sccm  
Example 2  
For a 0-10.0 l/min unit calibrated for Argon the flow at 5.0VDC would be  
10.0l/min. The K factor for Argon is 1.45. If the unit is used with Carbon  
Dioxide (K factor 0.74) then the flow rate 5.0VDC would be  
(0.74/1.45)10.0 = 5.10l/min  
The accuracy of readings using K factors is not as good as that achieved  
for the calibration gas. The accuracy obtained (typically ±3% for K factors  
similar to the calibration gas) depends on the gas being used and the flow  
rate.  
For a list of common K Factors see Section J.  
4. Changing The Flow Rate Set-Point (Using An External Voltage  
Source)  
The required flow rate is selected by adjusting the set-point voltage. The  
normal control signal voltage is 0-5VDC with 0VDC corresponding to zero  
flow and 5VDC being equivalent to the maximum rated flow of the unit.  
This input is linear and scaleable allowing different flow rates within the  
range of the unit to be selected. For example:  
For a flow range of 0-500sccm:  
A 5 VDC Input Signal would correspond to a flow rate of 500sccm  
If a flow rate of 300sccm were required then the set-point would  
be:  
(300 ÷ 500) × 5 = 3.0VDC  
M-4271/0707, pg. 15 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
If a gas other than the calibration gas is used then the adjusted maximum  
(full scale) flow for the unit should be calculated using the K Factor for  
that gas (see section C3 above).  
A zero or negative set-point voltage will cause the solenoid valve to close  
fully. Whilst closed, the valve is configured to withstand pressures up to 60  
psig (higher pressures on request).  
Caution: The flow controller valve will open if the pressure  
exceeds 60psig. For safety it is recommended that a separate  
positive shut-off valve is installed upstream of the controller.  
5. Changing The Flow Rate Set-Point – FMA3400/3400ST Series  
Only  
On the FMA 3400/3400ST Series the set-point may be input from an  
external source or be supplied internally.  
For an external set-point, dip switch 1 should be OFF and dip switch 2  
ON. See section C4 above for details of how to adjust the set-point using  
an external voltage source.  
For an internal set-point, dip switch 1 should be ON and dipswitch 2 OFF.  
Adjustment of the internal set-point is made by turning the coarse and fine  
set-point potentiometers on the front panel of the display, with the gas is  
flowing, until the desired flow rate is achieved.  
FMA 3400/3400ST Series Set-Point Potentiometers  
6. Power Save Mode.  
To improve valve performance and reliability over time, the FMA  
3200/3400 Series features a Power Save Mode. This is activated after a  
prolonged application of a zero or negative set-point. When a control  
voltage greater than 0 VDC is applied after the Power Save Mode has been  
initiated there may be a short delay (1-2 secs) before the valve actuates.  
M-4271/0707, pg. 16 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Zero Adjustments  
The zero should be checked as detailed in section C part 2. If an  
adjustment is needed the Zero Potentiometer should be carefully turned  
until the output (VDC) becomes zero.  
Caution: Do NOT adjust the Gain Potentiometer when adjusting  
the zero or the unit will need to be recalibrated.  
Making Zero Adjustments Using a Small Flathead Screwdriver  
Care should be taken to only make small adjustments to the zero  
potentiometer. If too much of an adjustment is made and difficulties are  
being experienced in achieving a zero reading then turn the potentiometer  
fully anti-clockwise and begin making small clockwise adjustments until a  
zero reading is obtained.  
8. Recalibration  
If recalibration is required please contact the Omega Engineering  
Customer Service Department.  
9. Changing the Calibration Gas – FMA 3400/3400ST Series Only  
The FMA 3400/3400ST Series may be calibrated for up to three gases.  
These gases, their corresponding flow ranges and accuracy specifications  
are detailed on the calibration certificate.  
The calibration gas required is determined by selecting the corresponding  
dip switch on the front panel of the display.  
M-4271/0707, pg. 17 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
FMA 3400/3400ST Series Dip Switches  
Dip switch 4 is allocated to the primary calibration gas.  
Dip switch 5 is allocated to the second calibration gas (if applicable).  
Dip switch 6 is allocated to the third calibration gas (if applicable).  
To select the gas, the dip switch should be turned ON. All other switches  
allocated to gases (i.e. 4, 5 or 6 except the required switch/gas) should be  
set to OFF.  
D.  
Maintenance and Product Care  
1. General  
Inlet filters should be periodically checked and cleaned or replaced as  
necessary.  
Regularly check all electrical and process connections for damage or  
deterioration.  
If the sensor is to be stored, keep both the inlet and outlet ports sealed.  
Do not allow any liquid or moisture to enter the sensor or damage will  
occur.  
2. Returning Units for Repair or Recalibration  
To return a unit for repair or recalibration please contact the Omega  
Engineering Customer Service Department. An Authorized Return (AR)  
number will then be issued. The AR number should then be noted on the  
outside of the package and on any correspondence. Further details may be  
found on page 25 of this manual.  
M-4271/0707, pg. 18 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
E.  
Specifications  
Series  
FMA 3200  
FMA 3400  
FMA 3200ST  
FMA 3400ST  
Accuracy  
(including linearity)  
±1.5% of  
Full Scale*  
±1.5% of  
Full Scale*  
Second and  
third gases  
±3.0% F.S.*  
±1.5% of  
Full Scale*  
±1.5% of  
Full Scale*  
Second and  
third gases  
±3.0% F.S.*  
Repeatability  
±0.25% Full Scale*  
Pressure Rating  
Pressure Sensitivity  
Temperature Rating  
150 psig (10.3 bar)  
500 psig (34.5 bar)  
±0.02% Full Scale* per psi (per 69 mbar)  
Operating Range: 5 to 55ºC  
Recommended Range (for best performance) : 10 to 40ºC  
Storage Range: 0 to 70ºC  
Temperature  
Sensitivity  
±0.15% F.S.* or less per ºC  
Valve  
Normally Closed  
Positive Shut-off up to 60psig (4 bar)  
Body Leak Integrity  
Wetted Materials  
1x10-7 sccs of He  
Aluminum  
304 Stainless Steel  
316 Stainless Steel  
303 Stainless Steel  
304 Stainless Steel  
316 Stainless Steel  
Epoxy  
O-Ring Material  
Fitting Material  
Viton®  
Choose from acetal, brass, or stainless steel  
Recommended  
Filtration  
20 microns or less  
Optional inline filters available  
Compatible gases  
Output Signal  
Clean, dry gases compatible with wetted materials  
0-5VDC, Impedance greater than 2.5 K  
0-5VDC, Integrated 2Mload  
External Set-point  
Signal  
Internal Set-point  
Signal  
N/A  
N/A  
Front Panel  
Adjustment  
N/A  
Front Panel  
Adjustment  
Warm-Up Time  
Less than 5 minutes  
Integrated Display  
3½ digit  
N/A  
3½ digit  
Typical Power  
Consumption  
Standard: 12 VDC @ 250 mA (12.5-15 VDC)  
“E” Suffix: 24 VDC @ 130 mA (22-25 VDC)  
Peak Power  
Consumption  
Standard: 12 VDC @ 400 mA (12.5-15 VDC)  
“E” Suffix: 24 VDC @ 260 mA (22-25 VDC)  
Electrical  
Connections  
Integrated 36” (92 mm) cable, terminated with:  
Standard: 6-pin Mini-DIN male (PS/2 Style)  
D1 Option: 9-pin D-Sub male  
D2 Option: 15-pin D-Sub male  
Certifications  
CE Approved  
89/336/EEC (EN 55011 & EN 50082-1)  
73/23/EEC Low Voltage Directive  
*Specifications from 10-100% of rated flow. Linearity is best fit straight line. All calibrations  
performed with air unless otherwise stated on calibration certificate.  
M-4271/0707, pg. 19 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F.  
Dimensions  
ALL DIMENSIONS IN INCHES (MILLIMETERS IN BRACKETS)  
FMA 3200/3200ST Series - 1/4” Stainless Fittings Shown  
M-4271/0707, pg. 20 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
FMA 3400/3400ST Series - 1/4” Stainless Fittings Shown  
M-4271/0707, pg. 21 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
G.  
Gas K Factors  
Chemical  
Symbol  
Gas  
K Factor  
Acetylene  
Air  
C2H2  
-
0.589  
1.000  
1.438  
0.260  
0.739  
1.000  
0.598  
0.330  
0.354  
0.385  
0.420  
0.460  
0.570  
1.458  
1.011  
1.440  
0.721  
1.443  
0.990  
1.000  
0.710  
0.991  
0.446  
0.383  
0.690  
1.437  
Argon  
Ar  
Butane  
C4H10  
CO2  
D2  
Carbon Dioxide  
Deuterium  
Ethylene  
Freon 11  
Freon 12  
Freon 13  
Freon 14  
Freon 22  
Germane  
Helium  
C2H4  
CCL3F  
CCL2F2  
CCLF3  
CF4  
CHCLF2  
GeH4  
He  
Hydrogen  
Krypton  
H2  
Kr  
Methane  
Neon  
CH4  
Ne  
Nitric Oxide  
Nitrogen  
Nitrous Oxide  
Oxygen  
NO  
N2  
N2O  
O2  
Ozone  
O3  
Propane  
Sulfur Dioxide  
Xenon  
C3H8  
SO2  
Xe  
These K Factors are given for reference only and are not intended as a recommendation of  
application suitability. Accuracy and response will be affected depending on the gas and  
flow range. Check the compatibility of all wetted materials before using any gas other than  
the calibration gas for the unit.  
M-4271/0707, pg. 22 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
H.  
Troubleshooting Guide  
Symptom  
Possible Cause  
Method of Correction  
Clicking noise from  
controller  
Unit in error mode  
Check there is sufficient  
pressure and that the flow  
path is not restricted or  
blocked  
No response  
Unit wired incorrectly  
Loose connection  
Check wiring is according to  
Section B5  
Check all connectors and  
wiring  
Damaged connector pins  
Blocked flow path  
Contact Omega Engineering  
Check flow path for  
obstructions.  
Piping leak before sensor  
Insufficient power  
Check all piping and  
connections.  
Check the power supply  
output and increase if  
necessary  
Output load resistance too low  
Flow too low for the unit  
Ensure the voltmeter or  
data acquisition system or  
display has an impedance of  
greater than of 2.5kohm  
Ensure that the flow being  
measured is within the  
capabilities of the unit  
Unit damaged or faulty  
Particles in flow path  
Contact Omega Engineering  
Inaccurate control  
Add filtration before the  
sensor.  
Flow path obscured  
Remove any debris or  
blockage in the flow path  
eg. PTFE tape.  
Unit calibrated for a different  
gas  
Check calibration certificate  
and apply a “K” Factor to  
readings if necessary.  
Gas composition is variable  
Fittings have been changed  
Contact Omega Engineering  
Replace the factory installed  
fittings  
Moisture in gas  
Ensure gas is clean and dry  
Insufficient warm-up period  
Allow the unit to warm-up  
for at least 5 minutes.  
Zero drift  
Verify the zero and adjust  
as necessary as explained in  
Section C  
The gain potentiometer has  
been adjusted  
Contact Omega Engineering  
Unit needs recalibration  
Flow too high for the unit  
Contact Omega Engineering  
Ensure that the flow being  
measured is within the  
capabilities of the unit  
M-4271/0707, pg. 23 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Symptom  
Possible Cause  
Method of Correction  
Inaccurate control  
Insufficient or varying power  
Check the power supply  
output and increase if  
necessary  
Insufficient pressure  
Varying pressure  
Ensure the pressure (an  
differential pressure) is high  
enough o operate the unit.  
Check the stability of the  
pressure regulation and  
improve if necessary.  
Ambient temperature too high  
or too low  
Place the unit in a suitable  
environment  
Output load resistance too low  
Ensure the voltmeter or  
data acquisition system or  
display has an impedance of  
greater than of 2.5kohm  
Gas temperature too high or  
too low  
Ensure the gas temperature  
is within the recommended  
operating range  
Unit damaged or faulty  
Contact Omega Engineering  
Problems with  
rezeroing  
Gas flow through unit not  
completely stopped  
Ensure there is no flow  
through the unit. The  
easiest way to do this is to  
plug both the inlet and  
outlet.  
Severe fluctuations in the  
ambient temperature e.g. unit  
in direct sunlight  
Carry out the rezero  
procedure in a stable  
environment  
Unstable power supply  
Check the stability and  
suitability of the power  
source  
Insufficient warm-up period  
Allow the unit to warm-up  
for at least 5 minutes.  
M-4271/0707, pg. 24 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
.
M-4271/0707, pg. 25 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
M-4271/0707, pg. 26 of 26  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Miele Refrigerator F 1471 User Manual
NEC Computer Monitor MultiSync 75F User Manual
Nokia Cell Phone 3310 User Manual
Nortel Networks Network Router 312865 A User Manual
Optoma Technology Projector TH1060P User Manual
Oricom Telephone Pro900 User Manual
Panasonic Camera Lens H PS45175 User Manual
Panasonic DVD Recorder DMR XW350 User Manual
Panasonic Network Card EZACT User Manual
Partner Tech Printer RP 300 H User Manual