National Instruments Switch NI 6115 6120 User Manual

DAQ  
NI 6115/6120 User Manual  
Multifunction I/O Devices for  
PCI/PXI/CompactPCI Bus Computers  
NI 6115/6120 User Manual  
November 2002 Edition  
Part Number 322812C-01  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Important Information  
Warranty  
The NI PCI-6115, NI PXI-6115, NI PCI-6120, and NI PXI-6120 devices are warranted against defects in materials and workmanship for a  
period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair  
or replace equipment that proves to be defective during the warranty period. This warranty includes parts and labor.  
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects  
in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National  
Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives  
notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be  
uninterrupted or error free.  
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before  
any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are  
covered by warranty.  
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical  
accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent  
editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected.  
In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.  
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF  
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMERS RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF  
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR  
DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY  
THEREOF. This limitation of the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including  
negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments  
shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover  
damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or  
maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire,  
flood, accident, actions of third parties, or other events outside reasonable control.  
Copyright  
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying,  
recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National  
Instruments Corporation.  
Trademarks  
CVI, DAQ-STC, LabVIEW, Measurement Studio, MITE, MXI, National Instruments, NI, ni.com, NI-DAQ, NI Developer  
Zone, and RTSIare trademarks of National Instruments Corporation.  
Product and company names mentioned herein are trademarks or trade names of their respective companies.  
Patents  
For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file  
on your CD, or ni.com/patents.  
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS  
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF  
RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN  
ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT  
INJURY TO A HUMAN.  
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE  
IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY,  
COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS  
AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND  
HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL  
DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR  
MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE  
HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD  
CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD  
NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID  
DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO  
PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS.  
BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING  
PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN  
COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL  
INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING  
THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE  
INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN,  
PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Compliance  
FCC/Canada Radio Frequency Interference Compliance  
Determining FCC Class  
The Federal Communications Commission (FCC) has rules to protect wireless communications from interference. The FCC  
places digital electronics into two classes. These classes are known as Class A (for use in industrial-commercial locations only)  
or Class B (for use in residential or commercial locations). Depending on where it is operated, this product could be subject to  
restrictions in the FCC rules. (In Canada, the Department of Communications (DOC), of Industry Canada, regulates wireless  
interference in much the same way.)  
Digital electronics emit weak signals during normal operation that can affect radio, television, or other wireless products. By  
examining the product you purchased, you can determine the FCC Class and therefore which of the two FCC/DOC Warnings  
apply in the following sections. (Some products may not be labeled at all for FCC; if so, the reader should then assume these are  
Class A devices.)  
FCC Class A products only display a simple warning statement of one paragraph in length regarding interference and undesired  
operation. Most of our products are FCC Class A. The FCC rules have restrictions regarding the locations where FCC Class A  
products can be operated.  
FCC Class B products display either a FCC ID code, starting with the letters EXN,  
or the FCC Class B compliance mark that appears as shown here on the right.  
Consult the FCC Web site at http://www.fcc.gov for more information.  
FCC/DOC Warnings  
This equipment generates and uses radio frequency energy and, if not installed and used in strict accordance with the instructions  
in this manual and the CE Marking Declaration of Conformity*, may cause interference to radio and television reception.  
Classification requirements are the same for the Federal Communications Commission (FCC) and the Canadian Department  
of Communications (DOC).  
Changes or modifications not expressly approved by National Instruments could void the user’s authority to operate the  
equipment under the FCC Rules.  
Class A  
Federal Communications Commission  
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC  
Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated  
in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and  
used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this  
equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct  
the interference at his own expense.  
Canadian Department of Communications  
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.  
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.  
Class B  
Federal Communications Commission  
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the  
FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.  
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the  
instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not  
occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can  
be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of  
the following measures:  
Reorient or relocate the receiving antenna.  
Increase the separation between the equipment and receiver.  
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.  
Consult the dealer or an experienced radio/TV technician for help.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Canadian Department of Communications  
This Class B digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.  
Cet appareil numérique de la classe B respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.  
Compliance to EU Directives  
Readers in the European Union (EU) must refer to the Manufacturer’s Declaration of Conformity (DoC) for information*  
pertaining to the CE Marking compliance scheme. The Manufacturer includes a DoC for most every hardware product except  
for those bought for OEMs, if also available from an original manufacturer that also markets in the EU, or where compliance is  
not required as for electrically benign apparatus or cables.  
To obtain the DoC for this product, click Declaration of Conformity at ni.com/hardref.nsf/. This Web site lists the DoCs  
by product family. Select the appropriate product family, followed by your product, and a link to the DoC appears in Adobe  
Acrobat format. Click the Acrobat icon to download or read the DoC.  
*
The CE Marking Declaration of Conformity will contain important supplementary information and instructions for the user  
or installer.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Conventions ...................................................................................................................xi  
NI-DAQ...........................................................................................................1-4  
National Instruments ADE Software...............................................................1-5  
Custom Cabling .............................................................................................................1-6  
Chapter 2  
Installing the Software...................................................................................................2-1  
Analog Output................................................................................................................3-5  
Analog Trigger...............................................................................................................3-5  
Antialiasing Filters.........................................................................................................3-8  
Phase-Locked Loop Circuit ...........................................................................................3-9  
Correlated Digital I/O ....................................................................................................3-10  
Timing Signal Routing...................................................................................................3-10  
© National Instruments Corporation  
vii  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Programmable Function Inputs....................................................................... 3-12  
GPCTR0_GATE Signal ................................................................... 4-33  
GPCTR0_OUT Signal...................................................................... 4-34  
GPCTR0_UP_DOWN Signal........................................................... 4-35  
GPCTR1_SOURCE Signal .............................................................. 4-35  
GPCTR1_GATE Signal ................................................................... 4-36  
GPCTR1_OUT Signal...................................................................... 4-36  
NI 6115/6120 User Manual  
viii  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
GPCTR1_UP_DOWN Signal...........................................................4-37  
Chapter 5  
Loading Stored Calibration Constants...........................................................................5-1  
Self-Calibration..............................................................................................................5-2  
External Calibration.......................................................................................................5-2  
Appendix A  
Specifications  
Appendix B  
Technical Support and Professional Services  
Glossary  
Index  
© National Instruments Corporation  
ix  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
About This Manual  
This manual describes the electrical and mechanical aspects of the  
NI 6115/6120 and contains information concerning its operation  
and programming.  
The NI 6115/6120 family includes the following devices:  
NI PCI-6115  
NI PXI-6115  
NI PCI-6120  
NI PXI-6120  
The NI 6115/6120 is a high-performance multifunction analog, digital, and  
timing I/O data acquisition (DAQ) device for PXI and PCI bus computers.  
Supported functions include analog input (AI), analog output (AO),  
digital I/O (DIO), and timing I/O (TIO).  
Conventions  
The following conventions appear in this manual:  
<>  
Angle brackets that contain numbers separated by an ellipsis represent a  
range of values associated with a bit or signal name—for example,  
DIO<3..0>.  
»
The » symbol leads you through nested menu items and dialog box options  
to a final action. The sequence File»Page Setup»Options directs you to  
pull down the File menu, select the Page Setup item, and select Options  
from the last dialog box.  
The symbol indicates that the following text applies only to a specific  
product, a specific operating system, or a specific software version.  
This icon denotes a note, which alerts you to important information.  
This icon denotes a caution, which advises you of precautions to take to  
avoid injury, data loss, or a system crash. When this symbol is marked on  
the product, refer to the Safety Information section of Chapter 1,  
Introduction, for precautions to take.  
© National Instruments Corporation  
xi  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
About This Manual  
bold  
Bold text denotes items that you must select or click in the software, such  
as menu items and dialog box options. Bold text also denotes parameter  
names and hardware labels.  
CompactPCI  
CompactPCI refers to the core specification defined by the PCI Industrial  
Computer Manufacturer’s Group (PICMG).  
italic  
Italic text denotes variables, emphasis, a cross reference, or an introduction  
to a key concept. This font also denotes text that is a placeholder for a word  
or value that you must supply.  
monospace  
Monospace text denotes text or characters that you should enter from the  
keyboard, sections of code, programming examples, and syntax examples.  
This font is also used for the proper names of disk drives, paths, directories,  
programs, subprograms, subroutines, device names, functions, operations,  
variables, filenames and extensions, and code excerpts.  
NI 6115/6120  
PCI  
This phrase refers to any device in the NI 6115/6120 family.  
Peripheral Component Interconnect—PCI is a high-performance  
expansion bus architecture originally developed by Intel to replace ISA  
and EISA.  
PXI  
A rugged, open system for modular instrumentation based on CompactPCI,  
with special mechanical, electrical, and software features. The PXI bus  
standard was originally developed by National Instruments in 1997, and is  
now managed by the PXI bus Systems Alliance.  
National Instruments Documentation  
The NI 6115/6120 User Manual is one piece of the documentation set for  
the DAQ system. You could have any of several types of documentation  
depending on the hardware and software in the system. Use the  
documentation you have as follows:  
DAQ Quick Start Guide—This guide describes how to install the DAQ  
software and hardware, and confirm that the DAQ device is operating  
properly. When using this guide, refer to the pinout diagram for the  
NI 6110/6111. The pinouts for the NI 6110/6111 and the NI 6115/6120  
are identical.  
DAQ hardware documentation—This documentation has detailed  
information about the DAQ hardware that plugs into or is connected to  
the computer. Use this documentation for hardware installation and  
NI 6115/6120 User Manual  
xii  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
About This Manual  
configuration instructions, specification information about the DAQ  
hardware, and application hints.  
Software documentation—You may have both application software  
and NI-DAQ documentation. NI application software includes  
LabVIEW and Measurement Studio. After you set up the hardware  
system, use either your application software documentation or the  
NI-DAQ documentation to help you write your application. If you  
have a large, complicated system, it is worthwhile to look through the  
software documentation before you configure the hardware.  
Accessory installation guides or manuals—If you are using accessory  
products, read the terminal block and cable assembly installation  
guides. They explain how to physically connect the relevant pieces  
of the system. Consult these guides when you are making the  
connections.  
Related Documentation  
The following documents contain information you may find helpful:  
DAQ Quick Start Guide, located at ni.com/manuals  
DAQ-STC Technical Reference Manual, located at ni.com/manuals  
NI Developer Zone tutorial, Field Wiring and Noise Considerations  
for Analog Signals, located at ni.com/zone  
NI-DAQ User Manual for PC Compatibles, located at  
ni.com/manuals  
NI-DAQ Function Reference Help. You can access this help file by  
clicking Start»Programs»National Instruments»NI-DAQ»  
NI-DAQ Help.  
PCI Local Bus Specification Revision 2.2  
PICMG 2.0 R3.0, CompactPCI Core Specification  
PXI Specification Revision 2.0, available from www.pxisa.org  
© National Instruments Corporation  
xiii  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
1
Introduction  
This chapter describes the NI 6115/6120, lists what you need to get started,  
describes the optional software and optional equipment, and explains how  
to unpack the device.  
About the NI 6115/6120  
Thank you for buying an NI 6115/6120. The NI 6115/6120 is a Plug and  
Play multifunction analog, digital, and timing I/O device for PXI and PCI  
bus computers. The NI 6115 features a 12-bit A/D converter (ADC) per  
channel with four simultaneously sampling analog inputs, and two 12-bit  
D/A converters (DACs) with voltage outputs. The NI 6120 features a 16-bit  
ADC per input channel and 16-bit DACs for output. Each device features  
eight lines of TTL-compatible correlated DIO, and two 24-bit  
counter/timers for TIO.  
The NI 6115/6120 is a DAQ device for PXI or the PCI bus. The device  
is software configured and calibrated, and completely switchless and  
jumperless. This feature is made possible by the NI MITE bus interface  
chip that connects the device to the PXI or PCI I/O bus. The MITE  
implements the PCI Local Bus Specification so that you can configure  
all the interrupts and base memory addresses with software.  
The NI 6115/6120 uses the NI data acquisition system timing controller  
(DAQ-STC) for time-related functions. The DAQ-STC consists of three  
timing groups that control AI, AO, and general-purpose counter/timer  
functions. These groups include a total of seven 24-bit and three 16-bit  
counters and a maximum timing resolution of 50 ns. The DAQ-STC makes  
possible such applications as buffered pulse generation and equivalent time  
sampling.  
The NI 6115/6120 uses the Real-Time System Integration (RTSI) bus to  
easily synchronize several measurement devices to a common trigger or  
timing event. The RTSI bus allows synchronization of the measurements.  
The RTSI bus consists of the RTSI bus interface and a ribbon cable to route  
timing and trigger signals between as many as five DAQ devices in the  
computer. If you are using the NI PXI-6115/6120 in a PXI chassis, RTSI  
lines, known as the PXI trigger bus, are part of the backplane. Therefore,  
© National Instruments Corporation  
1-1  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                   
Chapter 1  
Introduction  
you do not need the RTSI cable for system triggering and timing on the  
PXI. In addition, a phase-locked loop (PLL) circuit accomplishes the  
synchronization of multiple NI PXI-6115/6120 devices or other PXI  
devices which support PLL synchronization by allowing these devices to  
all lock to the same reference clock present on the PXI backplane. Refer to  
the Phase-Locked Loop Circuit section of Chapter 3, Hardware Overview,  
for more information.  
Detailed specifications of the NI 6115/6120 are in Appendix A,  
Specifications.  
Using PXI with CompactPCI  
The ability to use PXI-compatible products with standard CompactPCI  
products is an important feature of PXI Specification Revision 2.0. If you  
use a PXI-compatible plug-in device in a standard CompactPCI chassis,  
you are unable to use PXI-specific functions, but you can still use the basic  
plug-in device functions. For example, the RTSI interface on the  
NI PXI-6115/6120 is available in a PXI chassis, but not in a CompactPCI  
chassis.  
The CompactPCI specification permits vendors to develop sub-buses that  
coexist with the basic PCI interface on the CompactPCI bus. Compatible  
operation is not guaranteed between CompactPCI devices with different  
sub-buses nor between CompactPCI devices with sub-buses and PXI  
devices. The standard implementation for CompactPCI does not include  
these sub-buses. The NI PXI-6115/6120 works in any standard  
CompactPCI chassis adhering to PICMG CompactPCI 2.0 R3.0.  
PXI-specific features are implemented on the J2 connector of the  
CompactPCI bus. Table 1-1 lists the J2 pins used by the NI PXI-6115/6120.  
The PXI device is compatible with any CompactPCI chassis with a sub-bus  
that does not drive these lines. Even if the sub-bus is capable of driving  
these lines, the PXI device is still compatible as long as those pins on the  
sub-bus are disabled by default and are never enabled.  
Caution Damage can result if these lines are driven by the sub-bus.  
NI 6115/6120 User Manual  
1-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 1  
Introduction  
Table 1-1. NI PXI-6115/6120 J2 Pin Assignment  
PXI J2 Pin  
Number  
NI PXI-6115/6120 Signal  
PXI Pin Name  
RTSI<0..5>  
PXI Trigger<0..5>  
B16, A16, A17,  
A18, B18, C18  
RTSI 6  
Star  
D17  
E16  
RTSI Clock  
Reserved  
PXI Trigger 7  
LBL<0..12>  
C20, E20, A19,  
C19  
Reserved  
LBR<0..12>  
A21, C21, D21,  
E21, A20, B20,  
E15, A3, C3, D3,  
E3, A2, B2  
What You Need to Get Started  
To set up and use the NI 6115/6120, you need the following:  
A computer or a PXI/CompactPCI chassis and controller  
(hereafter referred to as the computer)  
At least one of the following devices:  
NI PCI-6115  
NI PXI-6115  
NI PCI-6120  
NI PXI-6120  
NI 6115/6120 User Manual  
NI-DAQ for PC compatibles  
(Optional) One of the following software packages and  
documentation:  
LabVIEW (Windows)  
Measurement Studio (Windows)  
VI Logger (Windows)  
© National Instruments Corporation  
1-3  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 1  
Introduction  
Software Programming Choices  
When programming National Instruments DAQ hardware, you can use an  
NI application development environment (ADE) or other ADEs. In either  
case, you use NI-DAQ.  
NI-DAQ  
NI-DAQ, which shipped with the NI 6115/6120, has an extensive library of  
functions that you can call from the ADE. These functions allow you to use  
all the features of the device.  
NI-DAQ carries out many of the complex interactions, such as  
programming interrupts, between the computer and the DAQ hardware.  
NI-DAQ maintains a consistent software interface among its different  
versions so that you can change platforms with minimal modifications  
to the code. Whether you are using LabVIEW, LabWindows/CVI,  
Measurement Studio, VI Logger, or other ADEs, your application uses  
NI-DAQ, as illustrated in Figure 1-1.  
NI 6115/6120 User Manual  
1-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 1  
Introduction  
LabVIEW,  
LabWindows/CVI,  
Measurement Studio,  
or VI Logger  
Conventional  
Programming Environment  
NI-DAQ  
Driver Software  
Personal  
Computer or  
Workstation  
DAQ  
Hardware  
Figure 1-1. The Relationship Among the Programming Environment,  
NI-DAQ, and the Hardware  
To download a free copy of the most recent version of NI-DAQ, click  
Download Software at ni.com.  
National Instruments ADE Software  
LabVIEW features interactive graphics, a state-of-the-art interface, and  
a powerful graphical programming language. The LabVIEW Data  
Acquisition VI Library, a series of virtual instruments for using LabVIEW  
with National Instruments DAQ hardware, is included with LabVIEW.  
LabWindows/CVI is a complete ANSI C ADE that features an interactive  
user interface, code generation tools, and the LabWindows/CVI Data  
Acquisition and Easy I/O libraries.  
Measurement Studio, which includes tools for Visual C++ and tools for  
Visual Basic, is a development suite that allows you to design test and  
measurement applications. For Visual Basic developers, Measurement  
Studio features a set of ActiveX controls for using National Instruments  
© National Instruments Corporation  
1-5  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 1  
Introduction  
DAQ hardware. These ActiveX controls provide a high-level programming  
interface for building virtual instruments (VIs). For Visual C++ developers,  
Measurement Studio offers a set of Visual C++ classes and tools to  
integrate those classes into Visual C++ applications. The ActiveX controls  
and classes are available with Measurement Studio and the NI-DAQ  
software.  
VI Logger is an easy-to-use yet flexible tool specifically designed for data  
logging applications. Using dialog windows, you can configure data  
logging tasks to easily acquire, log, view, and share your data. VI Logger  
does not require any programming; it is a stand-alone, configuration-based  
software program.  
Using LabVIEW, LabWindows/CVI, Measurement Studio, or VI Logger  
greatly reduces the development time for your data acquisition and control  
application.  
Optional Equipment  
NI offers a variety of products to use with the NI 6115/6120, including  
cables, connector blocks, and other accessories, as follows:  
Shielded cables and cable assemblies  
Connector blocks, shielded 50- and 68-pin screw terminals  
RTSI bus cables (PCI only)  
Low channel-count signal conditioning modules, devices, and  
accessories, including conditioning for strain gauges, resistance  
temperature detectors, and relays  
For more specific information about these products, refer to the NI catalog  
at ni.com/catalog.  
Custom Cabling  
NI offers cables and accessories to help you prototype your application or  
to use if you frequently change device interconnections.  
If you want to develop your own cable, however, adhere to the following  
guidelines for best results:  
For AI signals, use shielded twisted-pair wires for each AI pair for  
differential inputs. Tie the shield for each signal pair to the ground  
reference at the source.  
NI 6115/6120 User Manual  
1-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 1  
Introduction  
Route the analog lines separately from the digital lines.  
When using a cable shield, use separate shields for the analog and  
digital halves of the cable. Failure to do so results in noise coupling  
into the analog signals from transient digital signals.  
Mating connectors and a backshell kit for making custom 68-pin cables are  
available from NI.  
The parts in the following list are recommended for connectors that mate to  
the I/O connector on the device:  
Honda 68-position, solder cup, female connector  
Honda backshell  
Unpacking  
The NI 6115/6120 is shipped in an antistatic package to prevent  
electrostatic damage to the device. Electrostatic discharge (ESD) can  
damage several components on the device.  
Caution Never touch the exposed pins of connectors.  
To avoid such damage when handling the device, take the following  
precautions:  
Ground yourself using a grounding strap or by holding a grounded  
object.  
Touch the antistatic package to a metal part of the computer chassis  
before removing the device from the package.  
Remove the device from the package and inspect the device for  
loose components or any sign of damage. Notify NI if the device appears  
damaged in any way. Do not install a damaged device into the computer.  
Store the NI 6115/6120 in the antistatic envelope when not in use.  
© National Instruments Corporation  
1-7  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 1  
Introduction  
Safety Information  
The following section contains important safety information that you must  
follow when installing and using the product.  
Do not operate the product in a manner not specified in this document.  
Misuse of the product can result in a hazard. You can compromise the  
safety protection built into the product if the product is damaged in any  
way. If the product is damaged, return it to National Instruments for repair.  
Do not substitute parts or modify the product except as described in this  
document. Use the product only with the chassis, modules, accessories, and  
cables specified in the installation instructions. You must have all covers  
and filler panels installed during operation of the product.  
Do not operate the product in an explosive atmosphere or where there may  
be flammable gases or fumes. If you must operate the product in such an  
environment, it must be in a suitably rated enclosure.  
If you need to clean the product, use a soft, nonmetallic brush. The product  
must be completely dry and free from contaminants before you return it to  
service.  
Operate the product only at or below Pollution Degree 2. Pollution is  
foreign matter in a solid, liquid, or gaseous state that can reduce dielectric  
strength or surface resistivity. The following is a description of pollution  
degrees:  
Pollution Degree 1 means no pollution or only dry, nonconductive  
pollution occurs. The pollution has no influence.  
Pollution Degree 2 means that only nonconductive pollution occurs in  
most cases. Occasionally, however, a temporary conductivity caused  
by condensation must be expected.  
Pollution Degree 3 means that conductive pollution occurs, or dry,  
nonconductive pollution occurs that becomes conductive due to  
condensation.  
You must insulate signal connections for the maximum voltage for which  
the product is rated. Do not exceed the maximum ratings for the product.  
Do not install wiring while the product is live with electrical signals. Do not  
remove or add connector blocks when power is connected to the system.  
Avoid contact between your body and the connector block signal when hot  
swapping modules. Remove power from signal lines before connecting  
them to or disconnecting them from the product.  
NI 6115/6120 User Manual  
1-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 1  
Introduction  
Operate the product at or below the installation category1 marked on the  
hardware label. Measurement circuits are subjected to working voltages2  
and transient stresses (overvoltage) from the circuit to which they are  
connected during measurement or test. Installation categories establish  
standard impulse withstand voltage levels that commonly occur in  
electrical distribution systems. The following is a description of installation  
categories:  
Installation Category I is for measurements performed on circuits not  
directly connected to the electrical distribution system referred to as  
MAINS3 voltage. This category is for measurements of voltages from  
specially protected secondary circuits. Such voltage measurements  
include signal levels, special equipment, limited-energy parts of  
equipment, circuits powered by regulated low-voltage sources, and  
electronics.  
Installation Category II is for measurements performed on circuits  
directly connected to the electrical distribution system. This category  
refers to local-level electrical distribution, such as that provided by a  
standard wall outlet (for example, 115 V for U.S. or 230 V for Europe).  
Examples of Installation Category II are measurements performed on  
household appliances, portable tools, and similar products.  
Installation Category III is for measurements performed in the building  
installation at the distribution level. This category refers to  
measurements on hard-wired equipment such as equipment in fixed  
installations, distribution boards, and circuit breakers. Other examples  
are wiring, including cables, bus-bars, junction boxes, switches,  
socket-outlets in the fixed installation, and stationary motors with  
permanent connections to fixed installations.  
Installation Category IV is for measurements performed at the primary  
electrical supply installation (<1,000 V). Examples include electricity  
meters and measurements on primary overcurrent protection devices  
and on ripple control units.  
1
2
3
Installation categories, also referred to as measurement categories, are defined in electrical safety standard IEC 61010-1.  
Working voltage is the highest rms value of an AC or DC voltage that can occur across any particular insulation.  
MAINS is defined as a hazardous live electrical supply system that powers equipment. Suitably rated measuring circuits may  
be connected to the MAINS for measuring purposes.  
© National Instruments Corporation  
1-9  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
2
Installing and Configuring the  
NI 6115/6120  
This chapter explains how to install and configure the NI 6115/6120.  
Installing the Software  
Before you install the NI 6115/6120, complete the following steps to install  
the software:  
1. Install the ADE, such as LabVIEW or Measurement Studio, according  
to the instructions on the CD and the release notes.  
2. Install NI-DAQ according to the instructions on the CD and the  
DAQ Quick Start Guide included with the device. When using the  
DAQ Quick Start Guide, refer to the pinout for the NI 6110/6111,  
which is identical to the pinout for the NI 6115/6120.  
Note It is important to install NI-DAQ before installing the NI 6115/6120 to ensure that  
the device is properly detected.  
Installing the Hardware  
You can install the NI 6115/6120 in any available expansion slot in the  
computer. However, to achieve best noise performance, leave as much  
room as possible between the NI 6115/6120 and other devices and  
hardware.  
The following are general installation instructions, so consult the computer  
user manual or technical reference manual for specific instructions and  
warnings.  
NI PXI-6115/6120  
1. Power off and unplug the computer.  
2. Choose an unused PXI slot in the system. For maximum performance,  
the NI PXI-6115/6120 has an onboard DMA controller that you can  
use only if the device is installed in a slot that supports bus arbitration,  
© National Instruments Corporation  
2-1  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 2  
Installing and Configuring the NI 6115/6120  
or bus master devices. NI recommends installing the  
NI PXI-6115/6120 in such a slot.  
Note The PXI specification requires all slots to support bus master devices, but the  
CompactPCI specification does not. If you install in a CompactPCI non-master slot, you  
must disable the onboard DMA controller using software.  
until they go out before continuing the installation.  
4. Remove the filler panel for the slot you have chosen.  
5. Ground yourself using a grounding strap or by touching a grounded  
object. Follow the ESD protection precautions described in the  
Unpacking section of Chapter 1, Introduction.  
6. Remove the rubber front panel screw protectors.  
7. Insert the NI PXI-6115/6120 into a 5 V PXI slot. Use the  
injector/ejector handle to fully insert the device into the chassis.  
8. Screw the front panel of the NI PXI-6115/6120 to the front  
panel-mounting rail of the system.  
9. Visually verify the installation. Make sure the device is not touching  
other devices or components and is fully inserted in the slot.  
10. Plug in and power on the computer.  
The NI PXI-6115/6120 is now installed. You are now ready to configure  
the hardware and software.  
NI PCI-6115/6120  
1. Power off and unplug the computer.  
2. Remove the cover.  
wait until they go out before continuing the installation.  
4. Remove the expansion slot cover on the back panel of the computer.  
5. Ground yourself using a grounding strap or by touching a grounded  
object. Follow the ESD protection precautions described in the  
Unpacking section of Chapter 1, Introduction.  
6. Insert the NI PCI-6115/6120 into a PCI system slot. Gently rock the  
device to ease it into place. It may be a tight fit, but do not force the  
device into place.  
7. Screw the mounting bracket of the device to the back panel rail of the  
computer.  
NI 6115/6120 User Manual  
2-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 2  
Installing and Configuring the NI 6115/6120  
8. Replace the cover.  
9. Plug in and power on the computer.  
The NI PCI-6115/6120 is now installed. You are now ready to configure the  
hardware and software.  
Configuring the Device  
Because of the NI standard architecture for data acquisition and the PCI bus  
specification, the NI 6115/6120 is completely software configurable. Two  
types of configuration are performed on the NI 6115/6120: bus-related and  
data acquisition-related configuration.  
The NI PCI-6115/6120 is fully compatible with the industry-standard  
PCI Local Bus Specification Revision 2.2. This compatibility allows the  
PCI system to automatically perform all bus-related configurations with no  
user interaction. Bus-related configuration includes setting the device base  
memory address and interrupt channel.  
The NI PXI-6115/6120 is fully compatible with the industry-standard PXI  
Specification Revision 2.0. This allows the PXI/CompactPCI system to  
automatically perform all bus-related configurations with no user  
interaction. Bus-related configuration includes setting the device base  
memory address and interrupt channel.  
Data acquisition-related configuration, which you must perform, includes  
such settings as analog input coupling and range, and others. You can  
modify these settings using NI-DAQ or application-level software, such as  
LabVIEW and Measurement Studio.  
To configure the device in Measurement & Automation Explorer (MAX),  
refer to either the DAQ Quick Start Guide or to the NI-DAQ User Manual  
for PC Compatibles at ni.com/manuals. For operating system-specific  
installation and troubleshooting instructions, refer to ni.com/  
support/daq.  
© National Instruments Corporation  
2-3  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
3
Hardware Overview  
This chapter presents an overview of the hardware functions on the  
NI 6115/6120. Figures 3-1 and 3-2 provide block diagrams for the NI 6115  
and NI 6120, respectively.  
CH0+  
+
Anti-  
Aliasing  
Filter  
CH0  
12-Bit  
ADC  
CH0  
Latch  
AI CH0  
Mux  
CH0  
Amplifier  
12  
12  
Data (16)  
Data (16)  
Data (16)  
CH0–  
CH1+  
+
Anti-  
Aliasing  
Filter  
CH1  
12-Bit  
ADC  
AI CH1  
Mux  
CH1  
Latch  
CH1  
Amplifier  
CH1–  
CH2+  
+
Anti-  
Aliasing  
Filter  
CH2  
12-Bit  
ADC  
AI CH2  
Mux  
CH2  
Latch  
12  
12  
CH2  
Amplifier  
Control  
CH2–  
CH3+  
Generic  
Bus  
Interface  
PCI  
Bus  
Interface  
Mini  
MITE  
ADC  
FIFO  
Data (32)  
+
Anti-  
Aliasing  
Filter  
CH3  
12-Bit  
ADC  
Address/Data  
AI CH3  
Mux  
CH3  
Latch  
CH3  
Amplifier  
Data (16)  
AI Control  
CH3–  
Calibration  
Mux  
EEPROM  
IRQ  
DMA  
Analog  
Trigger  
Circuitry  
2
Trigger  
Trigger Level  
DACs  
Analog Input  
Timing/Control  
DMA/IRQ  
Trigger  
PFI / Trigger  
Analog  
Input  
Control  
EEPROM  
Control  
DMA  
Interface  
Bus  
Interface  
Counter/  
Timing I/O  
DAQ-STC  
Bus  
Interface  
DIO  
FIFO  
DAQ - STC  
Timing  
FPGA  
I/O  
Bus  
Interface  
Analog  
Output  
Control  
Analog Output  
Timing/Control  
RTSI Bus  
Interface  
DIO  
Control  
STC Digital I/O (8)  
FPGA Digital I/O (8)  
Digital I/O  
DIO  
MUX  
Digital I/O (8)  
AO Control  
Data (32)  
Data (12)  
DAC0  
DAC1  
DAC  
FIFO  
RTSI Bus  
Calibration  
DACs  
Figure 3-1. NI 6115 Block Diagram  
© National Instruments Corporation  
3-1  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 3  
Hardware Overview  
CH0+  
+
Anti-  
Aliasing  
Filter  
CH0  
16-Bit  
ADC  
AI CH0  
Mux  
CH0  
Amplifier  
16  
16  
CH0  
Latch  
Data (16)  
Data (16)  
Data (16)  
CH0–  
CH1+  
+
Anti-  
Aliasing  
Filter  
CH1  
16-Bit  
ADC  
AI CH1  
Mux  
CH1  
Latch  
CH1  
Amplifier  
CH1–  
CH2+  
+
Anti-  
Aliasing  
Filter  
CH2  
16-Bit  
ADC  
AI CH2  
Mux  
CH2  
Latch  
16  
16  
CH2  
Amplifier  
Control  
CH2–  
CH3+  
Generic  
Bus  
Interface  
PCI  
Bus  
Interface  
Mini  
MITE  
ADC  
FIFO  
Data (32)  
+
Anti-  
Aliasing  
Filter  
CH3  
16-Bit  
ADC  
Address/Data  
AI CH3  
Mux  
CH3  
Latch  
CH3  
Amplifier  
Data (16)  
CH3–  
Calibration  
Mux  
EEPROM  
IRQ  
DMA  
Analog  
Trigger  
Circuitry  
2
Trigger  
Trigger Level  
DACs  
Analog Input  
Timing/Control  
DMA/IRQ  
Trigger  
PFI / Trigger  
Analog  
Input  
Control  
EEPROM  
Control  
DMA  
Interface  
Bus  
Interface  
Counter/  
Timing I/O  
DAQ-STC  
Bus  
Interface  
DAQ - STC  
DIO  
FIFO  
Timing  
FPGA  
Analog  
Output  
Control  
I/O  
Bus  
Interface  
Analog Output  
Timing/Control  
RTSI Bus  
Interface  
STC Digital I/O (8)  
FPGA Digital I/O (8)  
DIO  
Control  
Digital I/O  
DIO  
MUX  
Digital I/O (8)  
AO Control  
Data (32)  
DAC  
FIFO  
DAC0  
DAC1  
Data (16)  
RTSI Bus  
Calibration  
DACs  
Figure 3-2. NI 6120 Block Diagram  
Analog Input  
Input Mode  
The NI 6115/6120 supports only differential (DIFF) input mode. For more  
information about DIFF input, refer to the Connecting Analog Input  
Signals section of Chapter 4, Connecting Signals, which contains diagrams  
showing the signal paths for DIFF input mode.  
Note The inputs are differential only in the sense that the ground loops are broken.  
The negative input is not intended to carry signals of interest; rather it provides a DC  
reference point for the positive input, which may be different than ground.  
NI 6115/6120 User Manual  
3-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 3  
Hardware Overview  
Input Polarity and Input Range  
The NI 6115/6120 has bipolar inputs only. Bipolar input means that the  
midpoint of the input voltage range is centered at zero volts.  
You can independently configure each channel for a different input voltage  
range.  
The software-programmable gain on this device increases its overall  
flexibility by matching the input signal ranges to those that the ADC can  
accommodate. It has ranges of 42 V, 20 V, 10 V, 5 V, 2 V, 1 V,  
500 mV, and 200 mV and is suited for a wide variety of signal levels.  
By choosing the optimal gain setting, you can maximize usage of the  
dynamic range of the ADC, which effectively increases input signal  
resolution. Table 3-1 shows the overall input range and precision according  
to the gain used.  
Caution The NI 6115/6120 is not designed for input voltages greater than 42 VDC.  
Input voltages greater than 42 VDC can damage the NI 6115/6120, any device connected  
to it, and the host computer. Overvoltage can also cause an electric shock hazard for the  
operator. NI is not liable for damage or injury resulting from such misuse.  
© National Instruments Corporation  
3-3  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 3  
Hardware Overview  
Table 3-1. Input Range and Measurement Precision  
Precision1  
Input Range  
6115 (12-Bit)  
6120 (16-Bit)  
–50 to +50 V2  
–20 to +20 V  
–10 to +10 V  
–5 to +5 V  
–2 to +2 V  
–1 to +1V  
24.4 mV  
9.77 mV  
4.88 mV  
2.44 mV  
977 µV  
244 µV  
97.7 µV  
1.53 mV  
610 µV  
305 µV  
153 µV  
61.0 µV  
30.5 µV  
15.3 µV  
6.10 µV  
–500 to +500 mV  
–200 to +200 mV  
1 The value of 1 least significant bit (LSB) of the ADC; that is, the voltage increment  
corresponding to a change of one count in the ADC count.  
2 Do not exceed 42 VDC maximum.  
Note: Refer to Appendix A, Specifications, for absolute maximum ratings.  
Considerations for Selecting Input Ranges  
The range you select depends on the expected range of the incoming signal.  
A large input range can accommodate a large signal variation but reduces  
the voltage resolution. Choosing a smaller input range improves the voltage  
resolution but may result in the input signal going out of range. For best  
results, match the input range as closely as possible to the expected range  
of the input signal.  
Input Coupling  
You can configure the NI 6115/6120 for either AC or DC input coupling  
on a per channel basis. Use AC coupling when the AC signal contains a  
large DC component. If you enable AC coupling, you remove the large DC  
offset for the input amplifier and amplify only the AC component. This  
configuration effectively uses the dynamic range of the ADC.  
The input impedance for the programmable gain instrumentation amplifier  
(PGIA) channels is 1 Mfor ranges 10 V and 10 kfor  
ranges > 10 V. This configuration provides an AC-coupled corner  
frequency of 2.34 Hz for ranges 10 V and 234 Hz for ranges > 10 V.  
NI 6115/6120 User Manual  
3-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 3  
Hardware Overview  
Analog Output  
The NI 6115/6120 supplies two channels of AO voltage at the I/O  
connector. The range is fixed at bipolar 10 V.  
The AO channels on the NI 6115 contain 12-bit DACs that are capable of  
4 MS/s for one channel or 2.5 MS/s for each of two channels. The NI 6120  
DACs are 16-bit, and they have the same AO capabilities as the NI 6115.  
Refer to Appendix A, Specifications, for more detailed information about  
the AO capabilities of the NI 6115/6120.  
Note The AO channels do not have analog or digital filtering hardware and do produce  
images in the frequency domain related to the update rate.  
The NI 6115/6120 includes high-density memory modules allowing for  
long waveform generations.  
Analog Trigger  
In addition to supporting internal software triggering and external digital  
triggering to initiate a DAQ sequence, these devices also support analog  
hardware triggering. You can configure the analog trigger circuitry to  
accept either a direct analog input from the PFI0/TRIG1 pin on the  
I/O connector or a post-gain signal from the output of the PGIA on any of  
the channels, as shown in Figure 3-3. The trigger-level range for the direct  
analog channel is 10 V with a resolution of 78 mV for the NI 6115 and  
4.88 mV for the NI 6120. The input impedance for the direct analog  
channel is 10 kΩ. When this direct analog channel is configured for AC  
coupling, the corner frequency is 159 Hz.  
The range for the post-PGIA trigger from a selected channel is the  
full-scale range of the selected channel with a resolution of that range  
divided by 256 for the NI 6115 and 4,096 for the NI 6120.  
Two trigger reference signals, lowValue and highValue, can then be  
independently set to achieve advanced triggering modes. Refer to  
Figures 3-3 through 3-8 for illustrations of these modes.  
Note The PFI0/TRIG1 pin is an analog input when configured as an analog trigger.  
Therefore, it is susceptible to crosstalk from adjacent pins, which can result in false  
triggering when the pin is left unconnected. To avoid false triggering, make sure the  
PFI0/TRIG1 pin is connected to a low-impedance signal source (less than 1 ksource  
impedance) if you plan to enable this input using software.  
© National Instruments Corporation  
3-5  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 3  
Hardware Overview  
Digital Data  
ADC  
ADC  
+
Analog  
Input  
CH0  
PGIA  
PGIA  
PGIA  
PGIA  
highValue  
Trigger  
DAC  
+
Analog  
Input  
CH1  
ADC  
+
Analog  
Input  
CH2  
Analog  
Trigger  
Circuit  
Mux  
DAQ-STC  
ADC  
+
Analog  
Input  
CH3  
AC Couple  
Trigger  
DAC  
10 k  
lowValue  
PFI0/TRIG1  
Figure 3-3. Analog Trigger Block Diagram for the NI 6115/6120  
In below-low-level analog triggering mode, the trigger is generated when  
the signal value is less than lowValue, as shown in Figure 3-4. HighValue  
is unused.  
lowValue  
Trigger  
Figure 3-4. Below-Low-Level Analog Triggering Mode  
In above-high-level analog triggering mode, the trigger is generated when  
the signal value is greater than highValue, as shown in Figure 3-5.  
LowValue is unused.  
NI 6115/6120 User Manual  
3-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 3  
Hardware Overview  
highValue  
Trigger  
Figure 3-5. Above-High-Level Analog Triggering Mode  
In inside-region analog triggering mode, the trigger is generated when the  
signal value is between the lowValue and the highValue, as Figure 3-6  
shows.  
highValue  
lowValue  
Trigger  
Figure 3-6. Inside-Region Analog Triggering Mode  
In high-hysteresis analog triggering mode, the trigger is generated when the  
signal value is greater than highValue, with the hysteresis specified by  
lowValue, as Figure 3-7 shows.  
highValue  
lowValue  
Trigger  
Figure 3-7. High-Hysteresis Analog Triggering Mode  
© National Instruments Corporation  
3-7  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 3  
Hardware Overview  
In low-hysteresis analog triggering mode, the trigger is generated when  
the signal value is less than lowValue, with the hysteresis specified by  
highValue, as Figure 3-8 shows.  
highValue  
lowValue  
Trigger  
Figure 3-8. Low-Hysteresis Analog Triggering Mode  
The analog trigger circuit generates an internal digital trigger based on the  
AI signal and the user-defined trigger levels. This digital trigger can be  
used by any of the timing sections of the DAQ-STC, including the AI, AO,  
and general-purpose counter/timer sections. For example, the AI section  
can be configured to acquire n scans after the AI signal crosses a specific  
threshold. As another example, the AO section can be configured to update  
its outputs whenever the AI signal crosses a specific threshold.  
Antialiasing Filters  
Each AI channel on the NI 6115/6120 is equipped with a programmable  
antialaising Bessel filter. On the NI 6115, you can program the filters to  
provide a third-order 50 kHz lowpass filter, a third-order 500 kHz lowpass  
filter, or a pass-through mode with no filtering. On the NI 6120, you can  
program the filters to provide a five-pole 100 kHz low-pass filter or  
pass-through. These Bessel filters are highly effective at reducing signal  
aliasing and are designed for use with software filters.  
Existing software algorithms alone provide good roll-off at the cut-off  
frequency as shown in Figure 3-9. However, aliasing can cause  
high-frequency harmonics to make it through passbands in the filter. By  
combining hardware and software filtering, it is possible to obtain both  
steep roll-off and clean filtering of high-frequency aliases.  
NI 6115/6120 User Manual  
3-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 3  
Hardware Overview  
fcutoff  
fNyquist  
4 fcutoff  
(2 fcutoff  
)
Response of Analog Hardware Filters  
Response of Software Filters  
Figure 3-9. Effects of Hardware and Software Filtering on Antialiasing  
Phase-Locked Loop Circuit  
NI PXI-6115/6120  
A phase-locked loop (PLL) circuit accomplishes the synchronization of  
multiple NI PXI-6115/6120 devices or other PXI devices which support  
PLL synchronization by allowing these devices to all lock to the same  
reference clock present on the PXI backplane. This circuit allows you to  
trigger input or output operations on different devices and ensures that  
samples occur at the same time.  
The PLL circuitry consists of a voltage-controlled crystal oscillator  
(VCXO) with a tuning range of 50 ppm. The VCXO generates the  
60 MHz master clock used onboard the NI PXI-6115/6120.  
The PLL locks to the 10 MHz oscillator line on the PXI backplane bus.  
A phase comparator running at 1 MHz compares the PXI Bus and VCXO  
clock. The loop filter then processes the error signal and outputs a control  
voltage for the VCXO. Figure 3-10 illustrates the block diagram for the  
NI PXI-6115/6120.  
Note This feature is not available on the NI PCI-6115/6120.  
The PLL circuit is automatically enabled when the NI 6115/6120 is  
powered up. No configuration steps are required in order to utilize PLL  
synchronization.  
© National Instruments Corporation  
3-9  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 3  
Hardware Overview  
60 MHz out  
synched to 10 MHZ  
backplane clock  
Phase Comp  
+
10 MHz  
Div/10  
Div/60  
Loop  
Filter  
VCXO  
Figure 3-10. PLL Block Diagram  
Correlated Digital I/O  
You can software-configure groups of individual lines for either input or  
output. The NI 6115/6120 includes a FIFO for buffered operation. This  
operation allows you to read/write an array of data, using either an internal  
or external clock source, at a maximum rate of 10 MHz. In addition, you  
can correlate DIO and AI/AO operations to the same clock. Refer to the  
Correlating DIO Signal Connections section of Chapter 4, Connecting  
Signals, for information on which signals you can use to clock DIO  
operation. At system startup and reset, the DIO ports are all  
high-impedance.  
The hardware up/down control for general-purpose counters 0 and 1 are  
connected onboard to DIO6 and DIO7, respectively. Thus, you can use  
DIO6 and DIO7 to control the general-purpose counters. The up/down  
control signals, GPCTR0_UP_DOWN and GPCTR1_UP_DOWN, are  
input only and do not affect the operation of the DIO lines.  
Timing Signal Routing  
The DAQ-STC provides a flexible interface for connecting timing signals  
to other devices or external circuitry. The NI 6115/6120 uses the RTSI bus  
to interconnect timing signals between devices, and it uses the  
programmable function input (PFI) pins on the I/O connector to connect the  
device to external circuitry. These connections are designed to enable the  
NI 6115/6120 to both control and be controlled by other devices and  
circuits.  
There are 13 timing signals internal to the DAQ-STC that can be controlled  
by an external source. These timing signals can also be controlled by  
signals generated by the DAQ-STC, and these selections are fully software  
NI 6115/6120 User Manual  
3-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 3  
Hardware Overview  
configurable. For example, Figure 3-11 shows the signal routing  
multiplexer for controlling the STARTSCAN signal.  
RTSI Trigger <0..6>  
STARTSCAN  
PFI<0..9>  
Scan Interval Counter TC  
GPCTR0_OUT  
Figure 3-11. STARTSCAN Signal Routing  
This figure shows that STARTSCAN can be generated from a number of  
sources, including the external signals RTSI<0..6> and PFI<0..9> and the  
internal signals Scan Interval Counter TC and GPCTR0_OUT.  
Many of these timing signals are also available as outputs on the RTSI pins,  
as indicated in the RTSI Triggers section later in this chapter, and on the  
PFI pins, as indicated in Chapter 4, Connecting Signals.  
© National Instruments Corporation  
3-11  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 3  
Hardware Overview  
Programmable Function Inputs  
The 10 PFIs are connected to the signal routing multiplexer for each timing  
signal, and software can select one of the PFIs as the external source for a  
given timing signal. It is important to note that any PFI can be used as an  
input by any timing signal and that multiple timing signals can  
simultaneously use the same PFI. This flexible routing scheme reduces the  
need to change physical connections to the I/O connector for different  
applications. You also can individually enable each PFI pin to output a  
specific internal timing signal. For example, if you need the UPDATE*  
signal as an output on the I/O connector, software can enable the output  
driver for the PFI5/UPDATE* pin.  
Device and RTSI Clocks  
Many functions performed by the NI 6115/6120 require a frequency  
timebase to generate the necessary timing signals for controlling A/D  
conversions, updates, or general-purpose signals at the I/O connector.  
The NI 6115/6120 can use either its internal 20 MHz timebase or a  
timebase received over the RTSI bus. In addition, if you configure the  
device to use the internal timebase, you can also program the device to drive  
its internal timebase over the RTSI bus to another device that is  
programmed to receive this timebase signal. This clock source, whether  
local or from the RTSI bus, is used directly by the device as the primary  
frequency source. The default configuration at startup is to use the internal  
timebase without driving the RTSI bus timebase signal. This timebase is  
software selectable.  
RTSI Triggers  
The seven RTSI trigger lines on the RTSI bus provide a very flexible  
interconnection scheme for any device sharing the RTSI bus. These  
bidirectional lines can drive any of eight timing signals onto the RTSI bus  
and can receive any of these timing signals. The RTSI trigger lines connect  
to other devices through the PXI bus on the PXI backplane or through a  
special ribbon cable that must be installed for PCI. Figure 3-12 shows the  
PCI signal connection scheme and Figure 3-13 shows the PXI connection  
scheme.  
NI 6115/6120 User Manual  
3-12  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                 
Chapter 3  
Hardware Overview  
In PCI, you can access all seven RTSI lines (RTSI<0..6>) through their  
RTSI cable. With the NI PXI-6115/6120, RTSI<0..5> connects to PXI  
Trigger <0..5>, respectively, through the NI PXI-6115/6120 backplane.  
In PXI, RTSI<6> connects to the PXI Star Trigger line, allowing the  
NI 6115/6120 to receive triggers from any Star Trigger controller plugged  
into slot 2 of the chassis. For more information on the Star Trigger, refer to  
the PXI Specification Revision 2.0.  
DAQ-STC  
TRIG1  
TRIG2  
CONVERT*  
UPDATE*  
WFTRIG  
GPCTR0_SOURCE  
GPCTR0_GATE  
Trigger  
7
GPCTR0_OUT  
STARTSCAN  
AIGATE  
SISOURCE  
UISOURSE  
GPCTR1_SOURCE  
GPCTR1_GATE  
RTSI_OSC (20 MHz)  
Clock  
Switch  
Figure 3-12. PCI RTSI Bus Signal Connection  
© National Instruments Corporation  
3-13  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 3  
Hardware Overview  
DAQ-STC  
TRIG1  
TRIG2  
CONVERT*  
UPDATE*  
PXI Star (6)  
WFTRIG  
GPCTR0_SOURCE  
GPCTR0_GATE  
GPCTR0_OUT  
STARTSCAN  
AIGATE  
PXI Trigger (0..5)  
6
SISOURCE  
UISOURSE  
GPCTR1_SOURCE  
GPCTR1_GATE  
RTSI_OSC (20 MHz)  
PXI Trigger (7)  
Switch  
Figure 3-13. PXI RTSI Bus Signal Connections  
Refer to the Connecting Timing Signals section of Chapter 4, Connecting  
Signals, for a description of the signals shown in Figures 3-12 and 3-13.  
NI 6115/6120 User Manual  
3-14  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
4
Connecting Signals  
This chapter describes how to connect input and output signals to the  
NI 6115/6120 using the device I/O connector.  
Table 4-1. I/O Connector Details  
Device with I/O  
Connector  
Number  
of Pins  
Cable for Connecting  
to 100-pin Accessories  
Cable for Connecting  
to 68-pin Accessories  
NI 6115,  
NI 6120  
68  
N/A  
SH6868 Shielded Cable,  
SH68-68-EP, R6868,  
SH68-68R1-EP  
I/O Connector  
Figure 4-1 shows the pin assignments for the 68-pin I/O connector on the  
NI 6115/6120. A signal description follows the connector pinouts.  
Caution Connections that exceed any of the maximum ratings of input or output signals  
on the NI 6115/6120 can damage the device and the computer. NI is not liable for any  
damage resulting from such signal connections. The Protection column of Tables 4-3, 4-4,  
and 4-5 show the maximum input ratings for each signal.  
© National Instruments Corporation  
4-1  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 4  
Connecting Signals  
ACH0–  
ACH1+  
ACH1GND  
ACH2–  
ACH3+  
ACH3GND  
NC  
34 68  
33 67  
32 66  
31 65  
30 64  
29 63  
28 62  
27 61  
26 60  
25 59  
24 58  
23 57  
22 56  
21 55  
20 54  
19 53  
18 52  
17 51  
16 50  
15 49  
14 48  
13 47  
12 46  
11 45  
10 44  
ACH0+  
ACH0GND  
ACH1–  
ACH2+  
ACH2GND  
ACH3–  
NC  
NC  
NC  
NC  
NC  
NC  
NC  
AOGND  
AOGND  
DGND  
DIO0  
DIO5  
DGND  
DIO2  
DIO7  
DIO3  
SCANCLK  
EXTSTROBE*  
DGND  
PFI2/CONVERT*  
PFI3/GPCTR1_SOURCE  
PFI4/GPCTR1_GATE  
GPCTR1_OUT  
DGND  
NC  
NC  
NC  
NC  
NC  
DAC0OUT  
DAC1OUT  
NC  
DIO4  
DGND  
DIO1  
DIO6  
DGND  
+5V OUTPUT  
DGND  
DGND  
PFI0/TRIG1  
PFI1/TRIG2  
DGND  
+5V OUTPUT  
DGND  
PFI5/UPDATE*  
PFI6/WFTRIG  
DGND  
9
8
7
6
5
4
3
2
1
43  
42  
41  
40  
39  
38  
37  
36  
35  
PFI7/STARTSCAN  
PFI8/GPCTR0_SOURCE  
DGND  
PFI9/GPCTR0_GATE  
GPCTR0_OUT  
FREQ_OUT  
DGND  
NC = No Connect  
Figure 4-1. I/O Connector Pin Assignment for the NI 6115/6120  
NI 6115/6120 User Manual  
4-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 4  
Connecting Signals  
I/O Connector Signal Descriptions  
Table 4-2. Signal Descriptions for I/O Connector Pins  
Signal Name  
Reference  
Direction  
Description  
ACH<0..3>GND  
Ground for Analog Input Channels 0 through 3—These  
pins are the bias current return point for  
pseudodifferential measurements.  
ACH<0..3>+  
ACH<0..3>–  
ACH<0..3>GND  
ACH<0..3>GND  
Input  
Input  
Analog Input Channels 0 through 3 (+)—These pins  
are routed to the (+) terminal of the respective channel  
amplifier and carry the input signal.  
Analog Input Channels 0 through 3 (–)—These pins are  
routed to the (–) terminal of the respective channel  
amplifier and are the DC reference for the (+) input  
signal of that channel.  
DAC0OUT  
DAC1OUT  
AOGND  
DGND  
AOGND  
AOGND  
Output  
Output  
Analog Channel 0 Output—This pin supplies the  
voltage output of AO channel 0.  
Analog Channel 1 Output—This pin supplies the  
voltage output of AO channel 1.  
Analog Output Ground—The AO voltages are  
referenced to this node.  
Digital Ground—This pin supplies the reference for the  
digital signals at the I/O connector as well as the  
+5 VDC supply.  
DIO<0..7>  
DGND  
Input  
Output  
Digital I/O Signals—DIO6 and 7 can control the  
up/down signal of general-purpose counters 0 and 1,  
respectively.  
+5V  
DGND  
DGND  
Output  
Output  
+5 VDC Source—These pins are fused for up to 1 A of  
+5 V supply. The fuse is self-resetting.  
SCANCLK  
Scan Clock—This pin pulses once for each A/D  
conversion when enabled. The low-to-high edge  
indicates when the input signal can be removed from  
the input or switched to another signal.  
EXTSTROBE*  
DGND  
Output  
External Strobe—This output can be toggled under  
software control to latch signals or trigger events on  
external devices.  
© National Instruments Corporation  
4-3  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                         
Chapter 4  
Connecting Signals  
Table 4-2. Signal Descriptions for I/O Connector Pins (Continued)  
Signal Name  
PFI0/TRIG1  
Reference  
Direction  
Description  
DGND  
Input  
Output  
PFI0/Trigger 1—As an input, this is either a PFI or the  
source for the hardware analog trigger. PFI signals are  
explained in the Connecting Timing Signals section  
later in this chapter. The hardware analog trigger is  
explained in the Analog Trigger section of Chapter 3,  
Hardware Overview. As an output, this is the TRIG1  
signal. In posttrigger DAQ sequences, a low-to-high  
transition indicates the initiation of the DAQ sequence.  
In pretrigger applications, a low-to-high transition  
indicates the initiation of the pretrigger conversions.  
PFI1/TRIG2  
DGND  
Input  
Output  
PFI1/Trigger 2—As an input, this is a PFI. As an  
output, this is the TRIG2 signal. In pretrigger  
applications, a low-to-high transition indicates the  
initiation of the posttrigger conversions. TRIG2 is not  
used in posttrigger applications.  
PFI2/CONVERT*  
DGND  
DGND  
DGND  
Input  
Output  
PFI2/Convert—As an input, this is a PFI. As an output,  
this is the CONVERT* signal. A high-to-low edge on  
CONVERT* indicates that an A/D conversion is  
occurring.  
PFI3/GPCTR1_SOURCE  
PFI4/GPCTR1_GATE  
Input  
Output  
PFI3/Counter 1 Source—As an input, this is a PFI. As  
an output, this is the GPCTR1_SOURCE signal. This  
signal reflects the actual source connected to the  
general-purpose counter 1.  
Input  
Output  
PFI4/Counter 1 Gate—As an input, this is a PFI. As an  
output, this is the GPCTR1_GATE signal. This signal  
reflects the actual gate signal connected to the  
general-purpose counter 1.  
GPCTR1_OUT  
PFI5/UPDATE*  
DGND  
DGND  
Output  
Counter 1 Output—This output is from the  
general-purpose counter 1 output.  
Input  
Output  
PFI5/Update—As an input, this is a PFI. As an output,  
this is the UPDATE* signal. A high-to-low edge on  
UPDATE* indicates that the AO primary group is being  
updated.  
PFI6/WFTRIG  
DGND  
DGND  
Input  
Output  
PFI6/Waveform Trigger—As an input, this is a PFI.  
As an output, this is the WFTRIG signal. In timed  
AO sequences, a low-to-high transition indicates the  
initiation of the waveform generation.  
PFI7/STARTSCAN  
Input  
Output  
PFI7/Start of Scan—As an input, this is a PFI. As an  
output, this is the STARTSCAN signal. This pin pulses  
once at the start of each AI scan in the interval scan.  
A low-to-high transition indicates the start of the scan.  
NI 6115/6120 User Manual  
4-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                 
Chapter 4  
Connecting Signals  
Table 4-2. Signal Descriptions for I/O Connector Pins (Continued)  
Signal Name  
Reference  
Direction  
Description  
PFI8/GPCTR0_SOURCE  
DGND  
Input  
Output  
PFI8/Counter 0 Source—As an input, this is a PFI. As  
an output, this is the GPCTR0_SOURCE signal. This  
signal reflects the actual source connected to the  
general-purpose counter 0.  
PFI9/GPCTR0_GATE  
DGND  
Input  
Output  
PFI9/Counter 0 Gate—As an input, this is a PFI. As an  
output, this is the GPCTR0_GATE signal. This signal  
reflects the actual gate signal connected to the  
general-purpose counter 0.  
GPCTR0_OUT  
FREQ_OUT  
DGND  
DGND  
Output1  
Output  
Counter 0 Output—This output is from the  
general-purpose counter 0 output.  
Frequency Output—This output is from the frequency  
generator output.  
1 The GPCTR0_OUT acts as an input when using external clock mode with correlated DIO.  
Table 4-3. Analog I/O Signal Summary for the NI 6115  
Signal  
Type and  
Direction  
Impedance  
Input/  
Output  
Protection  
(Volts)  
On/Off  
Rise  
Time  
(ns)  
Source  
(mA at V)  
Sink  
(mA at V)  
Signal Name  
Bias  
ACH<0..3>+  
AI  
1 Min  
parallel with  
100 pF1  
42 V  
300 pA  
or 10 kin  
parallel with  
40 pF2  
ACH<0..3>–  
AI  
10 nF to  
ACH<0..3>  
GND  
42 V  
300 pA  
DAC0OUT  
DAC1OUT  
AO  
AO  
50 Ω  
Short-circuit  
to ground  
5 at 10  
5 at 10  
5 at –10  
5 at –10  
50 Ω  
Short-circuit  
to ground  
1 Applies to range 10 V, impedance refers to ACH<0..3>.  
2 Applies to range > 10 V, impedance refers to ACH<0..3>.  
© National Instruments Corporation  
4-5  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                 
Chapter 4  
Connecting Signals  
Table 4-4. Analog I/O Summary for the NI 6120  
Signal  
Type and  
Direction  
Impedance  
Input/  
Output  
Protection  
(Volts)  
On/Off  
Rise  
Time  
(ns)  
Source  
(mA at V)  
Sink  
(mA at V)  
Signal Name  
Bias  
ACH<0..3>+  
AI  
100 GΩ  
to GND  
42 V to  
GND  
300 pA  
ACH<0..3>–  
AI  
AI  
100 GΩ  
to GND  
42 V to  
GND  
200 pA  
300 pA  
Differential Pair  
ACH<0..3>+ to  
ACH<0..3>–  
1 Min parallel  
with 100 pF1 or  
10k in parallel  
with 40 pF2  
DAC0OUT  
DAC1OUT  
AO  
AO  
50 Ω  
Short-circuit  
to ground  
5 at 10  
5 at 10  
5 at –10  
5 at –10  
50 Ω  
Short-circuit  
to ground  
1 Applies to range 10 V, impedance refers to ACH<0..3>.  
2 Applies to range > 10 V, impedance refers to ACH<0..3>.  
Table 4-5. Digital I/O Signal Summary  
Signal  
Type and  
Direction  
Impedance  
Input/  
Output  
Protection  
(Volts)  
On/Off  
Rise  
Time  
(ns)  
Source  
(mA at V)  
Sink  
(mA at V)  
Signal Name  
Bias  
VCC  
DO  
DIO  
DO  
DO  
0.1 Ω  
Short-circuit  
to ground  
1 A  
24 at 0.4  
5 at 0.4  
5 at 0.4  
5 at 0.4  
1.1  
1.5  
1.5  
1.5  
DIO<0..7>  
VCC +0.5  
13 at  
(VCC –0.4)  
50 kΩ  
pu  
SCANCLK  
EXTSTROBE*  
PFI0/TRIG1  
35  
3.5 at  
(VCC –0.4)  
50 kΩ  
pu  
3.5 at  
(VCC –0.4)  
50 kΩ  
pu  
AI  
DIO  
10 kΩ  
3.5 at  
(VCC –0.4)  
9 kΩ  
pu and  
10 kΩ  
pd  
VCC +0.5  
PFI1/TRIG2  
DIO  
DIO  
VCC +0.5  
VCC +0.5  
3.5 at  
(VCC –0.4)  
5 at 0.4  
5 at 0.4  
1.5  
1.5  
50 kΩ  
pu  
PFI2/CONVERT*  
3.5 at  
50 kΩ  
(VCC –0.4)  
pu  
NI 6115/6120 User Manual  
4-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                 
Chapter 4  
Connecting Signals  
Table 4-5. Digital I/O Signal Summary (Continued)  
Signal  
Type and  
Direction  
Impedance  
Input/  
Output  
Protection  
(Volts)  
On/Off  
Rise  
Time  
(ns)  
Source  
(mA at V)  
Sink  
(mA at V)  
Signal Name  
Bias  
PFI3/GPCTR1_SOURCE  
DIO  
DIO  
DO  
VCC +0.5  
VCC +0.5  
3.5 at  
(VCC –0.4)  
5 at 0.4  
5 at 0.4  
5 at 0.4  
5 at 0.4  
5 at 0.4  
5 at 0.4  
5 at 0.4  
5 at 0.4  
5 at 0.4  
5 at 0.4  
1.5  
50 kΩ  
pu  
PFI4/GPCTR1_GATE  
GPCTR1_OUT  
3.5 at  
(VCC –0.4)  
1.5  
1.5  
1.5  
1.5  
1.5  
1.5  
1.5  
1.5  
1.5  
50 kΩ  
pu  
3.5 at  
(VCC –0.4)  
50 kΩ  
pu  
PFI5/UPDATE*  
DIO  
DIO  
DIO  
DIO  
DIO  
DIO  
DO  
VCC +0.5  
VCC +0.5  
VCC +0.5  
VCC +0.5  
VCC +0.5  
VCC +0.5  
3.5 at  
(VCC –0.4)  
50 kΩ  
pu  
PFI6/WFTRIG  
3.5 at  
(VCC –0.4)  
50 kΩ  
pu  
PFI7/STARTSCAN  
PFI8/GPCTR0_SOURCE  
PFI9/GPCTR0_GATE  
GPCTR0_OUT  
3.5 at  
(VCC –0.4)  
50 kΩ  
pu  
3.5 at  
(VCC –0.4)  
50 kΩ  
pu  
3.5 at  
(VCC –0.4)  
50 kΩ  
pu  
3.5 at  
(VCC –0.4)  
50 kΩ  
pu  
FREQ_OUT  
3.5 at  
50 kΩ  
(VCC –0.4)  
pu  
pu = pull up; pd = pull down; DO = Digital Output  
The tolerance on the 50 kpull-up and pull-down resistors is very large. Actual value may range between 17 and 100 kΩ.  
Types of Signal Sources  
When making signal connections, you must first determine whether the  
signal sources are floating or ground-referenced. The following sections  
describe these two types of signals.  
Floating Signal Sources  
A floating signal source is not connected in any way to the building ground  
system but, rather, has an isolated ground-reference point. Some examples  
of floating signal sources are outputs of transformers, thermocouples,  
battery-powered devices, optical isolator outputs, and isolation amplifiers.  
An instrument or device that has an isolated output is a floating signal  
© National Instruments Corporation  
4-7  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                         
Chapter 4  
Connecting Signals  
source. You must tie the ground reference of a floating signal to the  
NI 6115/6120 AI ground to establish a local or onboard reference for the  
signal. Otherwise, the measured input signal varies as the source floats out  
of the common-mode input range.  
Ground-Referenced Signal Sources  
A ground-referenced signal source is connected in some way to the  
building system ground and is, therefore, already connected to a common  
ground point with respect to the NI 6115/6120, assuming that the computer  
is plugged into the same power system. Non-isolated outputs of  
instruments and devices that plug into the building power system fall into  
this category.  
The difference in ground potential between two instruments connected to  
the same building power system is typically between 1 and 100 mV but can  
be much higher if power distribution circuits are not properly connected.  
If a grounded signal source is improperly measured, this difference may  
appear as an error in the measurement. The connection instructions for  
grounded signal sources are designed to eliminate this ground potential  
difference from the measured signal.  
Connecting Analog Input Signals  
The NI 6115/6120 channels are configured as pseudodifferential inputs.  
The input signal of each channel, ACH<0..3>+, is tied to the positive input  
of its PGIA, and each reference signal, ACH<0..3>–, is tied to the negative  
input of its PGIA. The inputs are differential only in the sense that ground  
loops are broken. The reference signal, ACH<0..3>–, is not intended to  
carry signals of interest but only to provide a DC reference point for  
ACH<0..3>+ that may be different from ground.  
Pseudodifferential signal connections increase common-mode noise  
rejection. They also allow input signals to float within the common-mode  
limits of the PGIA.  
NI 6115/6120 User Manual  
4-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
               
Chapter 4  
Connecting Signals  
Connections for Ground-Referenced Signal Sources  
Figures 4-2 and 4-3 show how to connect a ground-referenced signal  
source to a channel on the NI 6115 and NI 6120, respectively.  
AC Coupling  
1 M*  
Common-Mode  
Choke  
ACH0+  
Instrumentation  
Amplifier  
Ground-  
Referenced  
Signal  
+
+
Vs  
100 pf*  
PGIA  
Source  
+
Measured  
Voltage  
Vm  
ACH0–  
Common-  
Mode  
Noise and  
Ground  
10 nf  
+
Vcm  
Potential  
*10 k40 pf for ranges > 10 V  
ACH0GND  
I/O Connector  
ACH0 Connections Shown  
Figure 4-2. Pseudodifferential Input Connections on the NI 6115  
for Ground-Referenced Signals  
© National Instruments Corporation  
4-9  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Connecting Signals  
AC Coupling  
Instrumentation  
Amplifier  
+
ACH0+  
Ground-  
Referenced  
Signal  
+
Vs  
PGIA  
100 pF*  
1 M*  
+
Measured  
Voltage  
Source  
Vm  
Common-  
Mode  
Noise and  
Ground  
ACH0–  
High-Frequency  
Common Mode Choke  
+
Vcm  
Potential  
ACH0GND  
*10 k40 pf for ranges > 10 V  
50 Ω  
0.1 µF  
Figure 4-3. Pseudodifferential Input Connections on the NI 6120  
for Ground-Referenced Signals  
With this type of connection, the PGIA rejects both the common-mode  
noise in the signal and the ground potential difference between the signal  
source and the device ground, shown as Vcm in Figures 4-2 and 4-3.  
NI 6115/6120 User Manual  
4-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 4  
Connecting Signals  
Connections for Nonreferenced or Floating Signal Sources  
Figures 4-4 and 4-5 show how to connect a floating signal source to a  
channel on the NI 6115 and NI 6120, respectively.  
AC Coupling  
Common-Mode  
Choke  
ACH0+  
Instrumentation  
Amplifier  
+
Floating  
Signal  
Source  
+
Vs  
100 pf*  
1 M*  
10 nf  
PGIA  
+
Measured  
Voltage  
Vm  
ACH0–  
Bias  
Current  
Return  
Paths  
Bias  
Resistor  
(see text)  
*10 k40 pf for ranges > 10 V  
ACH0GND  
I/O Connector  
ACH0 Connections Shown  
Figure 4-4. Differential Input Connections on the NI 6115  
for Nonreferenced Signals  
© National Instruments Corporation  
4-11  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Connecting Signals  
AC Coupling  
Instrumentation  
Amplifier  
+
ACH0+  
ACH0–  
+
Floating  
Signal  
Source  
Vs  
PGIA  
100 pF*  
1 M*  
+
Measured  
Voltage  
Vm  
Bias  
Current  
Return  
Paths  
High-Frequency  
Common Mode Choke  
Bias  
Resistor  
(see text)  
*10 k40 pf for ranges > 10 V  
50 Ω  
0.1 µF  
ACH0GND  
I/O Connector  
Figure 4-5. Differential Input Connections on the NI 6120 for Nonreferenced Signals  
Figures 4-4 and 4-5 show a bias resistor connected between ACH0– and the  
floating signal source ground. This resistor provides a return path for the  
200 pA bias current. A value of 10 kto 100 kis usually sufficient.  
If you do not use the resistor and the source is truly floating, the source is  
not likely to remain within the common-mode signal range of the PGIA,  
and the PGIA saturates, causing erroneous readings. You must reference  
the source to the respective channel ground.  
Common-mode rejection might be improved by using another bias resistor  
from the ACH0+ input to ACH0GND. This connection gives a slight  
measurement error due to the voltage divider formed with the output  
impedance of the floating source, but it also gives a more balanced input for  
better common-mode rejection.  
Common-Mode Signal Rejection Considerations  
Figures 4-2 and 4-3 show connections for signal sources that are already  
referenced to some ground point with respect to the NI 6115/6120. In  
theory, the PGIA can reject any voltage caused by ground-potential  
differences between the signal source and the device. In addition, with  
pseudodifferential input connections, the PGIA can reject common-mode  
noise pickup in the leads connecting the signal sources to the device.  
NI 6115/6120 User Manual  
4-12  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 4  
Connecting Signals  
Like any amplifier, the common-mode rejection ratio (CMRR) of the PGIA  
is limited at high frequency. This limitation has been compensated for in the  
design of the NI 6115/6120 by using a common-mode choke on each  
channel.  
NI 6115  
The purpose of the 10 nF capacitance on the ACH<0..3>– connection of the  
NI 6115 is to provide an impedance for this choke to work against at high  
frequency, thus improving the high-frequency CMRR. Depending on your  
application and the type of common noise at your source, further  
common-noise rejection might be gained by placing a 0.1 µF ceramic  
bypass capacitor between ACH– and ACH0GND.  
Working Voltage Range  
The PGIA operates normally by amplifying signals of interest while  
rejecting common-mode signals as long as the following three conditions  
are met:  
1. The common-mode voltage (Vcm), which is equivalent to subtracting  
ACH<0..3>GND from ACH<0..3>– and which is shown in Figure 4-2,  
must be less than 2.5 V. This Vcm is a constant for all range selections.  
2. The signal voltage (Vs), which is equivalent to subtracting  
ACH<0..3>– from ACH<0..3>+ and which is shown in Figure 4-2,  
must be less than or equal to the range selection of the given channel.  
If Vs is greater than the range selected, the signal clips and information  
is lost.  
3. The total working voltage of the positive input, which can be thought  
of as (Vcm + Vs) or simply as subtracting ACH<0..3>GND from  
ACH<0..3>+, must be less than 11 V for ranges 10 V or less than  
42 V for ranges > 10 V.  
If any of these conditions are exceeded, current limiters limit the input  
current to 20 mA maximum into any input until the fault condition is  
removed.  
Note All inputs are protected at up to 42 V.  
© National Instruments Corporation  
4-13  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 4  
Connecting Signals  
Connecting Analog Output Signals  
The AO signals are DAC0OUT, DAC1OUT, and AOGND.  
DAC0OUT is the voltage output signal for AO channel 0. DAC1OUT is the  
voltage output signal for AO channel 1.  
AOGND is the ground-reference signal for the AO channels. AOGND is a  
hard ground.  
Figure 4-6 shows how to connect AO signals to the NI 6115/6120.  
DAC0OUT  
Channel 0  
+
Load  
Load  
VOUT 0  
AOGND  
VOUT 1  
+
DAC1OUT  
Channel 1  
Analog Output Channels  
NI 6115/6120  
Figure 4-6. Analog Output Connections  
NI 6115/6120 User Manual  
4-14  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Connecting Signals  
Connecting Digital I/O Signals  
The DIO signals are DIO<0..7> and DGND. DIO<0..7> are the signals  
making up the DIO port, and DGND is the ground-reference signal for the  
DIO port. You can program groups of individual lines to be inputs or  
outputs.  
Caution Exceeding the maximum input voltage ratings, which are listed in Table 4-3, can  
damage the NI 6115/6120 and the computer. NI is not liable for any damage resulting from  
such signal connections.  
Figure 4-7 shows signal connections for three typical DIO applications.  
+5 V  
LED  
DIO<4..7>  
TTL Signal  
DIO<0..3>  
+5 V  
Switch  
DGND  
I/O Connector  
NI 6115/6120  
Figure 4-7. Digital I/O Connections  
© National Instruments Corporation  
4-15  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 4  
Connecting Signals  
Figure 4-7 shows DIO<0..3> configured for digital input and DIO<4..7>  
configured for digital output. Digital input applications include receiving  
TTL signals and sensing external device states such as the switch state  
shown in Figure 4-7. Digital output applications include sending TTL  
signals and driving external devices such as the LED shown  
in Figure 4-7.  
Correlating DIO Signal Connections  
You can correlate DIO and AI/AO operations to the same clock on the  
NI 6115/6120. You can use any of the following signals as the clock  
source:  
AI Scan Start  
AO Update  
GPCTR  
RTSI<0..5>  
External Clock  
Notes To use either of the GPCTR signals or the external clock to clock DIO operations,  
you must use one RTSI<0..5> pin.  
To use an external clock for correlated DIO, the clock must have input on the Counter 0  
output pin (GPCTR0_OUT). In this case, be sure that this counter is not used in any other  
operation.  
The following timing diagrams illustrate the use of these signals as clock  
sources. You can software-configure DIO operations for either rising or  
falling edge on whichever clock you choose as the source. Figure 4-8 shows  
any clock signal, in general, driving two separate groups of lines configured  
for digital input (DI) and DO. The DI operation is using the rising edge of  
the clock and the DO operation is using its falling edge.  
NI 6115/6120 User Manual  
4-16  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Connecting Signals  
CLK  
DI<7:4>  
DO<3:0>  
Figure 4-8. Clock Signal Driving DI and DO Signals  
Figure 4-9 shows a DIO operation driven by the AO Update signal on its  
rising edge.  
AO Update  
DIO  
Figure 4-9. Rising-Edge AO Update Signal Driving a DIO Signal  
© National Instruments Corporation  
4-17  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Connecting Signals  
Figure 4-10 shows a DIO operation driven by an RTSI clock signal on its  
falling edge.  
RTSI  
DIO  
Figure 4-10. Falling-Edge RTSI Clock Signal Driving a DIO Signal  
Power Connections  
Two pins on the I/O connector supply +5 V from the computer power  
supply using a self-resetting fuse. The fuse resets automatically within a  
few seconds after the overcurrent condition is removed. These pins are  
referenced to DGND and can be used to power external digital circuitry.  
The power rating is +4.65 to +5.25 VDC at 1 A.  
Caution Under no circumstances should you connect these +5 V power pins directly to  
analog or digital ground or to any other voltage source on the NI 6115/6120 or any other  
device. Doing so can damage the NI 6115/6120 and the computer. NI is not liable for  
damage resulting from such connections.  
Connecting Timing Signals  
Caution Exceeding the maximum input voltage ratings, which are listed in Table 4-3, can  
damage the NI 6115/6120 and the computer. NI is not liable for any damage resulting from  
such signal connections.  
All external control over the timing of the NI 6115/6120 is routed through  
the 10 PFIs, labeled PFI0 through PFI9. These signals are explained in  
detail in the Programmable Function Input Connections section. These  
PFIs are bidirectional; as outputs they are not programmable and reflect  
the state of many DAQ, waveform generation, and general-purpose timing  
signals. There are five other dedicated outputs for the remainder of the  
timing signals. As inputs, the PFI signals are programmable and can  
control DAQ, waveform generation, and general-purpose timing signals.  
NI 6115/6120 User Manual  
4-18  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                 
Connecting Signals  
The DAQ signals are explained in the DAQ Timing Connections section  
later in this chapter. The Waveform Generation Timing Connections section  
later in this chapter explains the waveform generation signals, and the  
General-Purpose Timing Signal Connections section later in this chapter  
explains the general-purpose timing signals.  
All digital timing connections are referenced to DGND. This reference is  
demonstrated in Figure 4-11, which shows how to connect an external  
TRIG1 source and an external STARTSCAN source to two PFI pins on the  
NI 6115/6120.  
PFI0/TRIG1  
PFI7/STARTSCAN  
TRIG1  
Source  
STARTSCAN  
Source  
DGND  
I/O Connector  
NI 6115/6120  
Figure 4-11. Timing I/O Connections  
© National Instruments Corporation  
4-19  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 4  
Connecting Signals  
Programmable Function Input Connections  
You can externally control 13 internal timing signals from the PFI pins.  
The source for each of these signals is software selectable from any PFI  
when you want external control. This flexible routing scheme reduces the  
need to change the physical wiring to the device I/O connector for different  
applications requiring alternative wiring.  
You can individually enable each of the PFI pins to output a specific  
internal timing signal. For example, if you need the STARTSCAN signal as  
an output on the I/O connector, software can turn on the output driver for  
the PFI7/STARTSCAN pin. Be careful not to drive a PFI signal externally  
when it is configured as an output.  
As an input, each PFI can be individually configured for edge or level  
detection and for polarity selection, as well. You can use the polarity  
selection for any of the timing signals, but the edge or level detection  
depends upon the particular timing signal being controlled. The detection  
requirements for each timing signal are listed within the section that  
discusses that individual signal.  
In edge-detection mode, the minimum pulse width required is 10 ns.  
This setting applies for both rising-edge and falling-edge polarity settings.  
Edge-detect mode does not have a maximum pulse-width requirement.  
In level-detection mode, the PFIs themselves do not impose a minimum or  
maximum pulse-width requirement, but the particular timing signal being  
controlled can impose limits. These requirements are listed later in this  
chapter.  
DAQ Timing Connections  
The DAQ timing signals are TRIG1, TRIG2, STARTSCAN, CONVERT*,  
AIGATE, SISOURCE, SCANCLK, and EXTSTROBE*.  
Posttriggered data acquisition allows you to view only data that is acquired  
after a trigger event is received. A typical posttriggered DAQ sequence is  
shown in Figure 4-12.  
Note On the NI 6115/6120, each STARTSCAN pulse initiates one CONVERT* pulse,  
which simultaneously samples all channels.  
NI 6115/6120 User Manual  
4-20  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                       
Chapter 4  
Connecting Signals  
TRIG1  
STARTSCAN  
CONVERT*  
Scan Counter  
4
3
2
1
0
Figure 4-12. Typical Posttriggered Acquisition  
Pretriggered data acquisition allows you to view data that is acquired before  
the trigger of interest in addition to data acquired after the trigger.  
Figure 4-13 shows a typical pretriggered DAQ sequence. The description  
for each signal shown in these figures appears later in this chapter.  
TRIG1  
TRIG2  
n/a  
STARTSCAN  
CONVERT*  
Scan Counter  
3
2
1
0
2
2
2
1
0
Figure 4-13. Typical Pretriggered Acquisition  
TRIG1 Signal  
Any PFI pin can receive as an input the TRIG1 signal, which is available  
as an output on the PFI0/TRIG1 pin.  
Refer to Figures 4-12 and 4-13 for the relationship of TRIG1 to the DAQ  
sequence.  
As an input, TRIG1 is configured in the edge-detection mode. You can  
select any PFI pin as the source for TRIG1 and configure the polarity  
selection for either rising or falling edge. The selected edge of TRIG1 starts  
the DAQ sequence for both posttriggered and pretriggered acquisitions.  
The NI 6115/6120 supports analog triggering on the PFI0/TRIG1 pin.  
© National Instruments Corporation  
4-21  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 4  
Connecting Signals  
Refer to Chapter 3, Hardware Overview, for more information on analog  
triggering.  
As an output, TRIG1 reflects the action that initiates a DAQ sequence even  
if another PFI is externally triggering the acquisition. The output is an  
active high pulse with a pulse width of 25 to 50 ns. This output is set to  
high-impedance at startup.  
Figures 4-14 and 4-15 show the timing requirements for TRIG1.  
tw  
Rising-Edge  
Polarity  
Falling-Edge  
Polarity  
tw = 10 ns minimum  
Figure 4-14. TRIG1 Input Signal Timing  
tw  
tw = 25 – 50 ns  
Figure 4-15. TRIG1 Output Signal Timing  
The device also uses TRIG1 to initiate pretriggered DAQ operations. In  
most pretriggered applications, TRIG1 is generated by a software trigger.  
Refer to the TRIG2 signal description for a complete description of the use  
of TRIG1 and TRIG2 in a pretriggered DAQ operation.  
TRIG2 Signal  
Any PFI pin can receive as an input the TRIG2 signal, which is available  
as an output on the PFI1/TRIG2 pin. Refer to Figure 4-13 for the  
relationship of TRIG2 to the DAQ sequence.  
As an input, TRIG2 is configured in the edge-detection mode. You can  
select any PFI pin as the source for TRIG2 and configure the polarity  
NI 6115/6120 User Manual  
4-22  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 4  
Connecting Signals  
selection for either rising or falling edge. The selected edge of TRIG2  
initiates the posttriggered phase of a pretriggered DAQ sequence. In  
pretriggered mode, the TRIG1 signal initiates the acquisition. The scan  
counter indicates the minimum number of scans before TRIG2 can be  
recognized. After the scan counter decrements to zero, it is loaded with the  
number of posttrigger scans to acquire while the acquisition continues.  
The device ignores TRIG2 if it is asserted prior to the scan counter  
decrementing to zero. After the selected edge of TRIG2 is received, the  
device acquires a fixed number of scans and the acquisition stops. This  
mode acquires data both before and after receiving TRIG2.  
As an output, TRIG2 reflects the posttrigger in a pretriggered DAQ  
sequence even if another PFI is externally triggering the acquisition.  
TRIG2 is not used in posttriggered data acquisition. The output is an active  
high pulse with a pulse width of 25 to 50 ns. This output is set to  
high-impedance at startup.  
Figures 4-16 and 4-17 show the timing requirements for TRIG2.  
tw  
Rising-Edge  
Polarity  
Falling-Edge  
Polarity  
tw = 10 ns minimum  
Figure 4-16. TRIG2 Input Signal Timing  
tw  
tw = 25 – 50 ns  
Figure 4-17. TRIG2 Output Signal Timing  
© National Instruments Corporation  
4-23  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Connecting Signals  
STARTSCAN Signal  
Any PFI pin can receive as an input the STARTSCAN signal, which is  
available as an output on the PFI7/STARTSCAN pin. Refer to Figures 4-18  
and 4-19 for the relationship of STARTSCAN to the DAQ sequence.  
As an input, STARTSCAN is configured in the edge-detection mode. You  
can select any PFI pin as the source for STARTSCAN and configure the  
polarity selection for either rising or falling edge. The selected edge of  
STARTSCAN initiates a scan. The sample interval counter starts if you  
select internally triggered CONVERT*.  
As an output, STARTSCAN reflects the actual start pulse that initiates a  
scan even if another PFI is externally triggering the starts. You have two  
output options. The first is an active high pulse with a pulse width of 25 to  
50 ns, which indicates the start of the scan. The second action is an active  
high pulse that terminates at the start of the last conversion in the scan,  
which indicates a scan in progress. STARTSCAN is deasserted toff after the  
last conversion in the scan is initiated. This output is set to high-impedance  
at startup.  
Figures 4-18 and 4-19 show the timing requirements for STARTSCAN.  
tw  
Rising-Edge  
Polarity  
Falling-Edge  
Polarity  
tw = 10 ns minimum  
Figure 4-18. STARTSCAN Input Signal Timing  
NI 6115/6120 User Manual  
4-24  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Connecting Signals  
tw  
tw = 25 – 50ns  
a. Start of Scan  
Start Pulse  
CONVERT*  
STARTSCAN  
toff  
t
off = 10 ns minimum  
b. Scan in Progress, Two Conversions per Scan  
Figure 4-19. STARTSCAN Output Signal Timing  
The CONVERT* pulses are masked off until the device generates  
STARTSCAN. If you are using internally generated conversions, the first  
CONVERT* appears when the onboard sample interval counter reaches  
zero. If you select an external CONVERT*, the first external pulse after  
STARTSCAN generates a conversion. STARTSCAN pulses should be  
separated by at least one scan period.  
A counter on the NI 6115/6120 internally generates STARTSCAN unless  
you select some external source. This counter is started by the TRIG1  
signal and is stopped by either software or the sample counter.  
Scans generated by either an internal or external STARTSCAN signal are  
inhibited unless they occur within a DAQ sequence. Scans occurring within  
a DAQ sequence may be gated by either the hardware signal AIGATE or  
the software command register gate.  
© National Instruments Corporation  
4-25  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 4  
Connecting Signals  
CONVERT* Signal  
Any PFI pin can receive as an input the CONVERT* signal, which is  
available as an output on the PFI2/CONVERT* pin.  
Refer to Figures 4-20 and 4-21 for the relationship of CONVERT* to the  
DAQ sequence.  
As an input, CONVERT* is configured in the edge-detection mode. You  
can select any PFI pin as the source for CONVERT* and configure the  
polarity selection for either rising or falling edge. The selected edge of  
CONVERT* initiates an A/D conversion.  
As an output, CONVERT* reflects the actual convert pulse that is  
connected to the ADC even if another PFI is externally generating the  
conversions. The output is an active low pulse with a pulse width of  
50 to 100 ns. This output is set to high-impedance at startup.  
Figures 4-20 and 4-21 show the input and output timing requirements for  
CONVERT*.  
tw  
Rising-Edge  
Polarity  
Falling-Edge  
Polarity  
tw = 10 ns minimum  
Figure 4-20. CONVERT* Input Signal Timing  
tw  
tw = 50 – 100 ns  
Figure 4-21. CONVERT* Output Signal Timing  
NI 6115/6120 User Manual  
4-26  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 4  
Connecting Signals  
The ADC switches to hold mode within 20 ns of the selected edge. This  
hold-mode delay time is a function of temperature and does not vary from  
one conversion to the next.  
The sample interval counter on the NI 6115/6120 device normally  
generates CONVERT* unless you select some external source. The counter  
is started by the STARTSCAN signal and continues to count down and  
reload itself until the scan is finished. It then reloads itself in preparation for  
the next STARTSCAN pulse.  
A/D conversions generated by either an internal or external CONVERT*  
signal are inhibited unless they occur within a DAQ sequence. Scans  
occurring within a DAQ sequence may be gated by either the hardware  
signal AIGATE or the software command register gate.  
AIGATE Signal  
Any PFI pin can receive as an input the AIGATE signal, which is not  
available as an output on the I/O connector. The AIGATE signal can mask  
off scans in a DAQ sequence. You can configure the PFI pin you select as  
the source for AIGATE in level-detection mode. You can configure the  
polarity selection for the PFI pin for either active high or active low.  
In the level-detection mode if AIGATE is active, the STARTSCAN signal  
is masked off and no scans can occur.  
AIGATE can neither stop a scan in progress nor continue a previously  
gated-off scan. Once a scan has started, AIGATE does not gate off  
conversions until the beginning of the next scan and, conversely, if  
conversions are being gated off, AIGATE does not gate them back on until  
the beginning of the next scan.  
SISOURCE Signal  
Any PFI pin can receive as an input the SISOURCE signal, which is not  
available as an output on the I/O connector. The onboard scan interval  
counter (SI) uses SISOURCE as a clock to time the generation of the  
STARTSCAN signal. You must configure the PFI pin you select as the  
source for SISOURCE in the level-detection mode. You can configure the  
polarity selection for the PFI pin for either active high or active low.  
The maximum allowed frequency is 20 MHz, with a minimum pulse width  
of 23 ns high or low. There is no minimum frequency limitation.  
© National Instruments Corporation  
4-27  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 4  
Connecting Signals  
Either the 20 MHz or 100 kHz internal timebase generates SISOURCE  
unless you select some external source. Figure 4-22 shows the timing  
requirements for SISOURCE.  
tp  
tw  
tw  
tp = 50 ns minimum  
tw = 23 ns minimum  
Figure 4-22. SISOURCE Signal Timing  
SCANCLK Signal  
SCANCLK is an output-only signal that generates a pulse with the leading  
edge occurring approximately 50 to 100 ns after an A/D conversion begins.  
The polarity of this output is software selectable but is typically configured  
so that a low-to-high leading edge can clock external AI multiplexers  
indicating when the input signal has been sampled and can be removed.  
This signal has a 450 ns pulse width and is software enabled.  
Note When using NI-DAQ, SCANCLK polarity is low-to-high and cannot be changed  
programmatically.  
Figure 4-23 shows the timing for SCANCLK.  
CONVERT*  
td  
SCANCLK  
tw  
td = 50 to 100 ns  
tw = 450 ns  
Figure 4-23. SCANCLK Signal Timing  
NI 6115/6120 User Manual  
4-28  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Connecting Signals  
EXTSTROBE* Signal  
EXTSTROBE* is an output-only signal that generates either a single pulse  
or a sequence of eight pulses in the hardware-strobe mode. An external  
device can use this signal to latch signals or to trigger events. In the  
single-pulse mode, software controls the level of EXTSTROBE*. A 10 µs  
and a 1.2 µs clock are available for generating a sequence of eight pulses in  
the hardware-strobe mode.  
Note EXTSTROBE* cannot be enabled through NI-DAQ.  
Figure 4-24 shows the timing for the hardware-strobe mode  
EXTSTROBE* signal.  
VOH  
VOL  
tw  
tw  
Figure 4-24. EXTSTROBE* Signal Timing  
Waveform Generation Timing Connections  
The AO group defined for the NI 6115/6120 is controlled by WFTRIG,  
UPDATE*, and UISOURCE.  
WFTRIG Signal  
Any PFI pin can receive as an input the WFTRIG signal, which is available  
as an output on the PFI6/WFTRIG pin.  
As an input, WFTRIG is configured in the edge-detection mode. You can  
select any PFI pin as the source for WFTRIG and configure the polarity  
selection for either rising or falling edge. The selected edge of WFTRIG  
starts the waveform generation for the DACs. If you select internally  
generated UPDATE*, the UI counter starts.  
As an output, WFTRIG reflects the trigger that initiates waveform  
generation, even if another PFI is externally triggering the waveform  
generation. The output is an active high pulse with a pulse width of  
25 to 50 ns. This output is set to high-impedance at startup.  
© National Instruments Corporation  
4-29  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
               
Chapter 4  
Connecting Signals  
Figures 4-25 and 4-26 show the timing requirements for WFTRIG.  
tw  
Rising-Edge  
Polarity  
Falling-Edge  
Polarity  
tw = 10 ns minimum  
Figure 4-25. WFTRIG Input Signal Timing  
tw  
tw = 25 – 50 ns  
Figure 4-26. WFTRIG Output Signal Timing  
UPDATE* Signal  
Any PFI pin can receive as an input the UPDATE* signal, which is  
available as an output on the PFI5/UPDATE* pin.  
As an input, UPDATE* is configured in the edge-detection mode. You can  
select any PFI pin as the source for UPDATE* and configure the polarity  
selection for either rising or falling edge. The selected edge of UPDATE*  
updates the outputs of the DACs. In order to use UPDATE*, you must set  
the DACs to posted-update mode.  
As an output, UPDATE* reflects the actual update pulse that is connected  
to the DACs, even if another PFI is externally generating the updates. The  
output is an active low pulse with a pulse width of 50 to 75 ns. This output  
is set to high-impedance at startup.  
NI 6115/6120 User Manual  
4-30  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 4  
Connecting Signals  
Figures 4-27 and 4-28 show the timing requirements for UPDATE*.  
tw  
Rising-Edge  
Polarity  
Falling-Edge  
Polarity  
tw = 10 ns minimum  
Figure 4-27. UPDATE* Input Signal Timing  
tw  
tw = 50 – 75 ns  
Figure 4-28. UPDATE* Output Signal Timing  
The DACs are updated within 100 ns of the leading edge. Separate the  
UPDATE* pulses with enough time that new data can be written to the DAC  
latches.  
The UI counter for the NI 6115/6120 normally generates UPDATE* unless  
you select some external source. The UI counter is started by the WFTRIG  
signal and can be stopped by software or the internal buffer counter (BC).  
D/A conversions generated by either an internal or external UPDATE*  
signal do not occur when gated by the software command register gate.  
UISOURCE Signal  
Any PFI pin can receive as an input the UISOURCE signal, which is not  
available as an output on the I/O connector. The UI counter uses  
UISOURCE as a clock to time the generation of the UPDATE* signal. You  
must configure the PFI pin you select as the source for UISOURCE in the  
level-detection mode. You can configure the polarity selection for the PFI  
pin for either active high or active low. Figure 4-29 shows the timing  
requirements for UISOURCE.  
© National Instruments Corporation  
4-31  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 4  
Connecting Signals  
tp  
tw  
tw  
tp = 50 ns minimum  
tw = 10 ns minimum  
Figure 4-29. UISOURCE Signal Timing  
The maximum allowed frequency is 20 MHz, with a minimum pulse width  
of 10 ns high or low. There is no minimum frequency limitation.  
Either the 20 MHz or 100 kHz internal timebase normally generates  
UISOURCE unless you select some external source.  
General-Purpose Timing Signal Connections  
The general-purpose timing signals are GPCTR0_SOURCE,  
GPCTR0_GATE, GPCTR0_OUT, GPCTR0_UP_DOWN,  
GPCTR1_SOURCE, GPCTR1_GATE, GPCTR1_OUT,  
GPCTR1_UP_DOWN, and FREQ_OUT.  
GPCTR0_SOURCE Signal  
Any PFI pin can receive as an input the GPCTR0_SOURCE signal, which  
is available as an output on the PFI8/GPCTR0_SOURCE pin.  
As an input, GPCTR0_SOURCE is configured in the edge-detection mode.  
You can select any PFI pin as the source for GPCTR0_SOURCE and  
configure the polarity selection for either rising or falling edge.  
As an output, GPCTR0_SOURCE reflects the actual clock connected to  
general-purpose counter 0, even if another PFI is externally inputting the  
source clock. This output is set to high-impedance at startup.  
NI 6115/6120 User Manual  
4-32  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 4  
Connecting Signals  
Figure 4-30 shows the timing requirements for GPCTR0_SOURCE.  
tp  
tw  
tw  
tp = 50 ns minimum  
tw = 10 ns minimum  
Figure 4-30. GPCTR0_SOURCE Signal Timing  
The maximum allowed frequency is 20 MHz, with a minimum pulse width  
of 10 ns high or low. There is no minimum frequency limitation.  
The 20 MHz or 100 kHz timebase normally generates GPCTR0_SOURCE  
unless you select some external source.  
GPCTR0_GATE Signal  
Any PFI pin can receive as an input the GPCTR0_GATE signal, which is  
available as an output on the PFI9/GPCTR0_GATE pin.  
As an input, GPCTR0_GATE is configured in the edge-detection mode.  
You can select any PFI pin as the source for GPCTR0_GATE and configure  
the polarity selection for either rising or falling edge. You can use the gate  
signal in a variety of applications to perform actions such as starting and  
stopping the counter, generating interrupts, and saving the counter contents.  
As an output, GPCTR0_GATE reflects the actual gate signal connected to  
general-purpose counter 0, even if another PFI is externally generating the  
gate. This output is set to high-impedance at startup.  
© National Instruments Corporation  
4-33  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 4  
Connecting Signals  
Figure 4-31 shows the timing requirements for GPCTR0_GATE.  
tw  
Rising-Edge  
Polarity  
Falling-Edge  
Polarity  
tw = 10 ns minimum  
Figure 4-31. GPCTR0_GATE Signal Timing in Edge-Detection Mode  
GPCTR0_OUT Signal  
This signal is available as an output on the GPCTR0_OUT pin.  
The GPCTR0_OUT signal reflects the terminal count (TC) of  
general-purpose counter 0. You have two software-selectable output  
options: pulse on TC and toggle output polarity on TC. The output polarity  
is software-selectable for both options. This output is set to  
high-impedance at startup. Figure 4-32 shows the timing of  
GPCTR0_OUT.  
Note When using external clocking mode with correlated DIO, this pin is used as an input  
for the external clock.  
TC  
GPCTR0_SOURCE  
GPCTR0_OUT  
(Pulse on TC)  
GPCTR0_OUT  
(Toggle output on TC)  
Figure 4-32. GPCTR0_OUT Signal Timing  
NI 6115/6120 User Manual  
4-34  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 4  
Connecting Signals  
GPCTR0_UP_DOWN Signal  
This signal can be received as an input on the DIO6 pin and is not available  
as an output on the I/O connector. The general-purpose counter 0 counts  
down when this pin is at a logic low and count up when it is at a logic high.  
You can disable this input so that software can control the up-down  
functionality and leave the DIO6 pin free for general use.  
GPCTR1_SOURCE Signal  
Any PFI pin can receive as an input the GPCTR1_SOURCE signal, which  
is available as an output on the PFI3/GPCTR1_SOURCE pin.  
As an input, GPCTR1_SOURCE is configured in the edge-detection mode.  
You can select any PFI pin as the source for GPCTR1_SOURCE and  
configure the polarity selection for either rising or falling edge.  
As an output, GPCTR1_SOURCE monitors the actual clock connected to  
general-purpose counter 1, even if another PFI is externally generating the  
source clock. This output is set to high-impedance at startup.  
Figure 4-33 shows the timing requirements for GPCTR1_SOURCE.  
tp  
tw  
tw  
tp = 50 ns minimum  
tw = 10 ns minimum  
Figure 4-33. GPCTR1_SOURCE Signal Timing  
The maximum allowed frequency is 20 MHz, with a minimum pulse width  
of 10 ns high or low. There is no minimum frequency limitation.  
The 20 MHz or 100 kHz timebase normally generates GPCTR1_SOURCE  
unless you select some external source.  
© National Instruments Corporation  
4-35  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 4  
Connecting Signals  
GPCTR1_GATE Signal  
Any PFI pin can receive as an input the GPCTR1_GATE signal, which is  
available as an output on the PFI4/GPCTR1_GATE pin.  
As an input, GPCTR1_GATE is configured in edge-detection mode. You  
can select any PFI pin as the source for GPCTR1_GATE and configure the  
polarity selection for either rising or falling edge. You can use the gate  
signal in a variety of applications to perform actions such as starting and  
stopping the counter, generating interrupts, and saving the counter contents.  
As an output, GPCTR1_GATE monitors the actual gate signal connected to  
general-purpose counter 1, even if another PFI externally generates the  
gate. This output is set to high-impedance at startup.  
Figure 4-34 shows the timing requirements for the GPCTR1_GATE signal.  
tw  
Rising-Edge  
Polarity  
Falling-Edge  
Polarity  
tw = 10 ns minimum  
Figure 4-34. GPCTR1_GATE Signal Timing in Edge-Detection Mode  
GPCTR1_OUT Signal  
This signal is available only as an output on the GPCTR1_OUT pin.  
The GPCTR1_OUT signal monitors the TC device general-purpose  
counter 1. You have two software-selectable output options: pulse on TC  
and toggle output polarity on TC. The output polarity is software selectable  
for both options. This output is set to high-impedance at startup.  
NI 6115/6120 User Manual  
4-36  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 4  
Connecting Signals  
Figure 4-35 shows the timing requirements for GPCTR1_OUT.  
TC  
GPCTR1_SOURCE  
GPCTR1_OUT  
(Pulse on TC)  
GPCTR1_OUT  
(Toggle output on TC)  
Figure 4-35. GPCTR1_OUT Signal Timing  
GPCTR1_UP_DOWN Signal  
This signal can be received as an input on the DIO7 pin and is not available  
as an output on the I/O connector. General-purpose counter 1 counts down  
when this pin is at a logic low and counts up at a logic high. This input can  
be disabled so that software can control the up-down functionality and  
leave the DIO7 pin free for general use. Figure 4-36 shows the timing  
requirements for the GATE and SOURCE input signals and the timing  
specifications for the OUT output signals.  
tsc  
tsp  
tsp  
VIH  
SOURCE  
VIL  
tgsu  
tgh  
VIH  
GATE  
VIL  
tgw  
tout  
VOH  
OUT  
VOL  
tsc  
tsp  
tgsu  
tgh  
tgw  
tout  
Source Clock Period  
Source Pulse Width  
Gate Setup Time  
Gate Hold Time  
Gate Pulse Width  
Output Delay Time  
50 ns minimum  
23 ns minimum  
10 ns minimum  
0 ns minimum  
10 ns minimum  
80 ns maximum  
Figure 4-36. GPCTR Timing Summary  
© National Instruments Corporation  
4-37  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 4  
Connecting Signals  
The GATE and OUT signal transitions shown in Figure 4-36 are referenced  
to the rising edge of the SOURCE signal. This timing diagram assumes that  
the counters are programmed to count rising edges. The same timing  
diagram, but with the source signal inverted and referenced to the falling  
edge of the source signal, would apply when the counter is programmed to  
count falling edges.  
The GATE input timing parameters are referenced to the signal at the  
SOURCE input or to one of the internally generated signals on the  
NI 6115/6120. Figure 4-36 shows the GATE signal referenced to the rising  
edge of a source signal. The gate must be valid (either high or low) for at  
least 10 ns before the rising or falling edge of a source signal for the gate to  
take effect at that source edge, as shown by tgsu and tgh in Figure 4-36. The  
gate signal is not required to be held after the active edge of the source  
signal.  
If you use an internal timebase clock, the gate signal cannot be  
synchronized with the clock. In this case, gates applied close to a source  
edge take effect either on that source edge or on the next one. This  
arrangement results in an uncertainty of one source clock period with  
respect to unsynchronized gating sources.  
The OUT output timing parameters are referenced to the signal at the  
SOURCE input or to one of the internally generated clock signals on the  
NI 6115/6120. Figure 4-36 shows the OUT signal referenced to the rising  
edge of a source signal. Any OUT signal state changes occur within 80 ns  
after the rising or falling edge of the source signal.  
FREQ_OUT Signal  
This signal is available only as an output on the FREQ_OUT pin. The  
frequency generator for the NI 6115/6120 outputs the FREQ_OUT pin.  
The frequency generator is a 4-bit counter that can divide its input clock by  
the numbers 1 through 16. The input clock of the frequency generator is  
software selectable from the internal 10 MHz and 100 kHz timebases. The  
output polarity is software selectable. This output is set to high-impedance  
at startup.  
NI 6115/6120 User Manual  
4-38  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 4  
Connecting Signals  
Field Wiring Considerations  
Environmental noise can seriously affect the measurement accuracy of the  
NI 6115/6120 if you do not take proper care when running signal wires  
between signal sources and the device. The following recommendations  
apply mainly to AI signal routing, although they also apply to signal  
routing in general.  
Minimize noise pickup and maximize measurement accuracy by taking the  
following precautions:  
Use differential AI connections to reject common-mode noise.  
Use individually shielded, twisted-pair wires to connect AI signals to  
the device. With this type of wire, the signals attached to the ACH+ and  
ACH– inputs are twisted together and then covered with a shield. You  
then connect this shield only at one point to the signal source ground.  
This kind of connection is required for signals traveling through areas  
with large magnetic fields or high electromagnetic interference.  
Route signals to the device carefully. Keep cabling away from noise  
sources. The most common noise source in a PCI DAQ system is the  
video monitor. Separate the monitor from the analog signals as far as  
possible.  
The following recommendations apply for all signal connections to the  
NI 6115/6120:  
Separate the NI 6115/6120 signal lines from high-current or  
high-voltage lines. These lines can induce currents in or voltages on  
the NI 6115/6120 signal lines if they run in parallel paths at a close  
distance. To reduce the magnetic coupling between lines, separate  
them by a reasonable distance if they run in parallel, or run the lines at  
right angles to each other.  
Do not run signal lines through conduits that also contain power lines.  
Protect signal lines from magnetic fields caused by electric motors,  
welding equipment, breakers, or transformers by running them through  
special metal conduits.  
For more information, refer to the NI Developer Zone tutorial, Field Wiring  
and Noise Consideration for Analog Signals, available at ni.com/zone.  
© National Instruments Corporation  
4-39  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
5
Calibration  
This chapter discusses the calibration procedures for the NI 6115/6120.  
NI-DAQ includes calibration functions for performing all of the steps in the  
calibration process.  
Calibration refers to the process of minimizing measurement and output  
voltage errors by making small circuit adjustments. On the NI 6115/6120,  
these adjustments take the form of writing values to onboard calibration  
DACs (CalDACs).  
Some form of device calibration is required for most applications. If you do  
not calibrate the device, the signals and measurements could have very  
large offset, gain, and linearity errors.  
Three levels of calibration are available to you and are described in this  
chapter. The first level is the fastest, easiest, and least accurate; whereas the  
last level is the slowest, most difficult, and most accurate.  
Loading Stored Calibration Constants  
The NI 6115/6120 is factory calibrated before shipment at approximately  
25 °C to the levels indicated in Appendix A, Specifications. The associated  
calibration constants—the values that were written to the CalDACs to  
achieve calibration in the factory—are stored in the onboard nonvolatile  
memory (EEPROM). Because the CalDACs have no memory capability,  
they do not retain calibration information when the device is unpowered.  
Loading calibration constants refers to the process of loading the CalDACs  
with the values stored in the EEPROM. NI-DAQ determines when this is  
necessary and does it automatically.  
In the EEPROM, there is a user-modifiable calibration area in addition to  
the permanent factory calibration area. Hence, you can load the CalDACs  
with values either from the original factory calibration or from a calibration  
that you subsequently performed.  
© National Instruments Corporation  
5-1  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 5  
Calibration  
This method of calibration is not very accurate because it does not take into  
account the fact that the device measurement and output voltage errors can  
vary with time and temperature. It is better to self-calibrate when the device  
is installed in the environment in which it is used.  
Self-Calibration  
The NI 6115/6120 can measure and correct for almost all of its  
calibration-related errors without any external signal connections. NI-DAQ  
software provides a self-calibration method. This self-calibration process,  
which generally takes two to five minutes, is the preferred method of  
assuring accuracy in your application. Initiate self-calibration to minimize  
the effects of any offset and gain drifts, particularly those caused by  
warming.  
Immediately after self-calibration, the only significant residual calibration  
error could be gain error due to time or temperature drift of the onboard  
voltage reference. This error is addressed by external calibration, which is  
discussed in the following section. If you are interested primarily in relative  
measurements, you can ignore a small amount of gain error, and  
self-calibration should be sufficient.  
External Calibration  
The NI 6115/6120 has an onboard calibration reference to ensure the  
accuracy of self-calibration. Its specifications are listed in Appendix A,  
Specifications. The reference voltage is measured at the factory and stored  
in the EEPROM for subsequent self-calibrations. This voltage is stable  
enough for most applications, but if you are using the device at an extreme  
temperature or if the onboard reference has not been measured for a year or  
more, you may wish to externally calibrate the device.  
An external calibration refers to calibrating the device with a known  
external reference rather than relying on the onboard reference.  
Redetermining the value of the onboard reference is part of this process and  
the results can be saved in the EEPROM, so you should not have to perform  
an external calibration very often. You can externally calibrate the device  
by calling the NI-DAQ calibration function.  
NI 6115/6120 User Manual  
5-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 5  
Calibration  
To externally calibrate your device, be sure to use a very accurate external  
reference. The reference should be several times more accurate than the  
device itself.  
For a detailed calibration procedure for the NI 6115/6120, click Manual  
Calibration Procedures at ni.com/calibration.  
© National Instruments Corporation  
5-3  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
A
Specifications  
This appendix lists the specifications of the NI 6115/6120. These  
specifications are typical at 25 °C unless otherwise noted.  
Analog Input  
Input Characteristics  
Number of channels ............................... 4 pseudodifferential  
Type of ADC  
Resolution  
NI 6115.................................... 12 bits, 1 in 4,096  
NI 6120.................................... 16 bits, 1 in 65,536  
Pipeline  
NI 6115.................................... 4  
NI 6120.................................... 0  
Sampling rate  
Maximum  
NI 6115.................................... 10 million S/s  
NI 6120.................................... 800 kS/s  
Minimum  
NI 6115.................................... 20 kS/s  
NI 6120.................................... No minimum  
Input impedance  
ACH+ to ACH–  
Range 10 V......................... 1 Min parallel with 100 pF  
Range > 10 V......................... 10 kin parallel with 40 pF  
ACH– to ACHGND  
NI 6115.................................... 10 nF  
NI 6120.................................... 100 GΩ  
© National Instruments Corporation  
A-1  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Appendix A  
Specifications  
ACH+ to ACHGND  
NI 6115.....................................100 GΩ  
NI 6120.....................................100 GΩ  
Input bias current.................................... 300 pA  
Input offset current ................................. 200 pA  
Input coupling.........................................DC/AC  
Max working voltage for all analog input channels  
Positive input (ACH+)..................... 11 V for ranges < 10 V;  
42 V for ranges 10 V  
Negative input (ACH–) ................... 2.5 V  
Overvoltage protection ........................... 42 V  
Input current during  
overvoltage conditions............................ 20 mA max  
Input FIFO size....................................... 16 or 32 MS  
Data transfers..........................................DMA, interrupts,  
programmed I/O  
DMA modes ...........................................Scatter-gather  
DC Transfer Characteristics  
INL  
NI 6115............................................ 0.35 LSB typ, 1 LSB max  
NI 6120............................................2.5 LSB max  
DNL  
NI 6115............................................ 0.25 LSB typ, 1 LSB max  
NI 6120............................................0.75 LSB typ, no missing codes  
Offset, gain error  
NI 6115............................................Refer to Table A-1  
NI 6120............................................Refer to Table A-2  
NI 6115/6120 User Manual  
A-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix A  
Specifications  
Table A-1. NI 6115 Analog Input DC Accuracy Information  
Absolute Accuracy  
Relative Accuracy  
Nominal  
Range  
(V)  
Noise +  
Absolute  
% of Reading  
Quantization (mV)  
Single  
Resolution (mV)  
Single  
Temp  
Drift  
(%/°C)  
Accuracy at  
Full Scale  
( mV)  
Full  
Scale  
Offset  
(mV)  
24 Hours  
1 Year  
0.348  
0.273  
0.028  
0.018  
0.038  
0.045  
0.060  
0.105  
Pt.  
Averaged  
Pt.  
Averaged  
50  
20  
10  
5
0.346  
0.271  
0.026  
0.016  
0.036  
0.043  
0.058  
0.103  
33  
13  
42  
3.6  
0.0229  
0.0229  
0.0004  
0.0004  
0.0004  
0.0004  
0.0004  
0.0004  
211.0  
69.4  
10.22  
4.61  
2.26  
1.23  
0.71  
0.39  
48  
4.8  
17  
1.4  
19  
1.9  
6.7  
8.3  
4.2  
1.8  
1.1  
0.69  
0.43  
0.72  
10  
1.0  
3.4  
0.36  
4.8  
2.1  
1.2  
0.80  
0.51  
0.48  
2
1.3  
0.16  
0.21  
1
0.68  
0.35  
0.15  
0.09  
0.12  
0.5  
0.2  
0.061  
0.039  
0.080  
0.051  
Table A-2. NI 6120 Analog Input DC Accuracy Information1  
Absolute Accuracy  
Relative Accuracy  
Nominal  
Range  
(V)  
Noise + Quantization  
(µV)  
Absolute  
Accuracy  
at Full  
Scale  
( mV)  
% of Reading  
24  
Resolution (µV)  
Single  
Temp  
Drift  
(%/°C)  
Full  
Scale  
Offset  
Single  
Hours  
0.157  
0.139  
0.033  
0.036  
0.039  
0.077  
0.100  
0.123  
1 Year  
(µV)2,3  
Pt.  
Averaged  
503.5  
201.4  
100.7  
50.4  
Pt.  
Averaged  
663.0  
265.2  
132.6  
66.3  
50  
20  
10  
5
0.159  
0.141  
0.034  
0.37  
7,799.2  
3,120.7  
1,561.1  
781.4  
344.0  
264.4  
178.8  
90.8  
5,621.9  
2,248.7  
1,124.4  
562.2  
0.0106  
0.0106  
0.0006  
0.0006  
0.0006  
0.0006  
0.0006  
0.0006  
86.92  
31.11  
4.94  
2.61  
1.15  
1.05  
0.69  
0.35  
6,629.9  
2652.0  
1326.0  
663.0  
265.2  
180.8  
180.8  
144.7  
2
0.041  
0.079  
0.102  
0.125  
224.9  
20.1  
26.5  
1
150.0  
13.7  
18.1  
0.5  
0.2  
144.3  
13.7  
18.1  
112.8  
11.0  
14.5  
1 Accuracies are valid for measurements following an internal calibration. Averaged numbers assume dithering and  
averaging of 100 single-channel readings. Measurement accuracies are listed for operational temperatures within 1 °C  
of internal calibration temperature and 10 °C of external or factory-calibration temperature.  
2 The offset might degrade by 3 LSB with filter enabled.  
3 The offset might degrade by 1 LSB when sampling above 500 kS/s.  
© National Instruments Corporation  
A-3  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Appendix A  
Specifications  
Dynamic Characteristics  
Interchannel skew...................................1 ns typ  
Analog filters  
Number  
NI 6115.....................................2  
NI 6120.....................................1  
Type  
NI 6115.....................................3-pole Bessel  
NI 6120.....................................5-pole Bessel  
Frequency  
NI 6115.....................................50 and 500 kHz  
(software-enabled)  
NI 6120.....................................100 kHz  
(software-enabled)  
Crosstalk .................................................–80 dB, DC to 100 kHz  
NI 6115/6120 User Manual  
A-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix A  
Specifications  
Table A-3. NI 6115 Analog Input Dynamic Characteristics  
Bandwidth  
(MHz)1  
SFDR Typ  
(dB)2  
SFDR Max  
(dB)  
CMRR  
System Noise  
4
Input Range  
50 V  
(dB)3  
(LSBrms  
0.35  
0.45  
0.35  
0.35  
0.45  
0.60  
0.80  
1.3  
)
5.5  
4.4  
7.2  
4.8  
4.8  
4.4  
4.4  
4.1  
78  
78  
81  
81  
85  
85  
85  
81  
70  
70  
75  
75  
75  
75  
75  
70  
34  
20 V  
40  
10 V  
46  
5 V  
52  
2 V  
60  
1 V  
66  
500 mV  
200 mV  
70  
72  
1 –3 dB frequency for input amplitude at 96% of the input range (–0.3 dB)  
2 Measured at 100 kHz with twelfth-order bandpass filter after signal source  
3 DC to 60 Hz  
4 LSBrms, including quantization  
Table A-4. NI 6120 Analog Input Dynamic Characteristics  
Bandwidth  
(MHz)1  
SFDR Typ  
(dB)2  
SFDR Max  
(dB)3  
CMRR  
(dB)4  
System Noise  
5
Input Range  
50 V  
(LSBrms  
)
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
95  
96  
95  
95  
96  
94  
90  
85  
90  
90  
90  
90  
90  
90  
85  
80  
60  
68  
1.2  
20 V  
1.2  
10 V  
76  
1.2  
5 V  
82  
1.5  
2 V  
90  
1.7  
1 V  
95  
2.0  
500 mV  
200 mV  
100  
105  
2.2  
2.8  
1 –3 dB frequency for input amplitude at 10% of the input range (–20 dB)  
2 Measured at 100 kHz with twelfth-order bandpass filter after signal source  
3 100% production tested at 100 kHz  
4 DC to 60 Hz  
5 LSBrms, not including quantization  
© National Instruments Corporation  
A-5  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
Specifications  
Full-Scale (–0.3 dB) Input Amplitude  
74  
68  
62  
56  
50  
44  
38  
10 V  
5 V  
2 V  
1 V  
0.5 V  
0.2 V  
0.1  
1.0  
10.0  
Frequency (MHz)  
Figure A-1. NI 6115 Total Harmonic Distortion Plus Noise (THD + N)  
NI 6115/6120 User Manual  
A-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
Specifications  
Full-Scale (–0.3 dB) Input Amplitude  
85  
83  
81  
79  
77  
75  
5 V  
2 V  
1 V  
0.5 V  
10 V  
0.2 V  
1
10  
100  
Frequency (kHz)  
Figure A-2. NI 6120 Total Harmonic Distortion Plus Noise (THD + N)  
© National Instruments Corporation  
A-7  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
Specifications  
High-Voltage Ranges, only 10 V Input Amplitude  
74  
68  
62  
56  
50  
44  
38  
50 V  
20 V  
0.1  
1.0  
10.0  
Frequency (MHz)  
Figure A-3. NI 6115 High-Voltage THD + N  
NI 6115/6120 User Manual  
A-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
Specifications  
High-Voltage Ranges, only 10 V Input Amplitude  
85.0  
84.5  
84.0  
83.5  
83.0  
82.5  
82.0  
81.5  
81.0  
80.5  
80.0  
50 V  
20 V  
1
10  
100  
Frequency (kHz)  
Figure A-4. NI 6120 High-Voltage THD + N  
© National Instruments Corporation  
A-9  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
Specifications  
With Filters, Full-Scale Input for Range of 1 V  
71.6  
70.4  
69.2  
68  
50 kHz  
500 kHz  
66.8  
65.6  
64.4  
63.2  
62  
10  
100  
1000  
Frequency (kHz)  
Figure A-5. NI 6115 THD + N with Filters  
NI 6115/6120 User Manual  
A-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
Specifications  
With Filters, Full-Scale Input for Range of 1 V  
85  
84  
83  
82  
81  
80  
79  
78  
77  
76  
75  
1
10  
100  
Frequency (kHz)  
Figure A-6. NI 6120 THD + N with Filters  
Stability  
Recommended warm-up time ................ 15 min  
Offset temperature coefficient  
Pregain  
NI 6115.................................... 12 µV/°C  
NI 6120.................................... 1.5 µV/°C  
Postgain  
NI 6115.................................... 64 µV/°C  
NI 6120.................................... 2.1 LSB/°C  
Gain temperature coefficient  
NI 6115 ........................................... 21.3 ppm/°C  
NI 6120 ........................................... 22.2 ppm/°C  
© National Instruments Corporation  
A-11  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix A  
Specifications  
Onboard calibration reference  
Level................................................5.000 V ( 2.5 mV)  
(actual value stored in EEPROM)  
Temperature coefficient................... 4.1 ppm/°C max  
Long-term stability.......................... 6 ppm/ 1,000 h  
Analog Output  
Output Characteristics  
Number of channels................................2 voltage  
Resolution  
NI 6115............................................12 bits, 1 in 4,096  
NI 6120............................................16 bits, 1 in 65,536  
Max update rate  
1 channel..........................................4 MS/s, system dependent  
2 channel..........................................2.5 MS/s, system dependent  
Output buffer size ...................................16 or 32 MS  
Data transfers..........................................DMA, interrupts,  
programmed I/O  
DMA modes ...........................................Scatter-gather  
DC Transfer Characteristics  
INL  
NI 6115............................................ 0.5 LSB typ, 2 LSB max  
NI 6120............................................ 0.35 LSB typ, 1 LSB max  
DNL  
NI 6115............................................ 0.25 LSB typ, 1 LSB max  
NI 6120............................................ 0.2 LSB typ, 1 LSB max  
Offset, gain error  
NI 6115............................................Refer to Table A-5  
NI 6120............................................Refer to Table A-6  
NI 6115/6120 User Manual  
A-12  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Appendix A  
Specifications  
Table A-5. NI 6115 Analog Output DC Accuracy Information  
Absolute Accuracy  
Relative Accuracy  
Nominal  
Range at  
Full Scale  
(V)  
% of Reading  
Temp  
Drift  
(%/°C)  
Absolute  
Acc. at Full  
Scale (mV)  
Offset  
(mV)  
Theoretical Resolution  
(mV)  
24 Hrs  
90 Days  
1 Year  
10  
0.0437  
0.0445  
0.0454  
8.9  
0.0006  
13.5  
4.88  
Table A-6. NI 6120 Analog Output DC Accuracy Information  
Absolute Accuracy  
Relative Accuracy  
Nominal  
Range at  
Full Scale  
(V)  
% of Reading  
Temp  
Drift  
(%/°C)  
Absolute  
Acc. at Full  
Scale (mV)  
Offset  
(mV)  
Theoretical Resolution  
(µV)  
24 Hrs  
90 Days  
1 Year  
10  
0.0512  
0.0520  
0.0529  
1.9  
0.0006  
6.7  
305.2  
Voltage Output  
Ranges.................................................... 10 V  
Output coupling...................................... DC  
Output impedance .................................. 50 5%  
Current drive .......................................... 5 mA  
Output stability....................................... Any passive load  
Protection .............................................. Short-circuit to ground  
Power-on output voltage  
(before software loads calibration values)  
NI 6115 ........................................... 400 mV  
NI 6120 ........................................... 80 mV  
Initial power-up glitch  
Magnitude ....................................... 2 V  
Duration .......................................... 200 ms  
© National Instruments Corporation  
A-13  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Appendix A  
Specifications  
Dynamic Characteristics  
Slew rate  
NI 6115............................................300 V/µs  
NI 6120............................................15 V/µs  
Noise  
NI 6115............................................600 µVrms, DC to 5 MHz  
NI 6120............................................100 µVrms, DC to 1 MHz  
Glitch energy at midscale transition  
NI 6115............................................ 30 mV for 1 µs  
NI 6120............................................ 10 mV for 1 µs  
Settling time  
NI 6115............................................300 ns to 0.01%  
NI 6120............................................4 µs to 1 LSB  
Stability  
Offset temperature coefficient  
NI 6115............................................ 35 µV/°C  
NI 6120............................................ 35 µV/°C  
Gain temperature coefficient  
NI 6115............................................ 56.9 ppm/°C  
NI 6120............................................ 6.5 ppm/°C  
Onboard calibration reference  
Level................................................5.000 V ( 2.5 mV)  
(actual value stored in EEPROM)  
Temperature coefficient................... 0.9 ppm/°C max  
Long-term stability.......................... 6 ppm/ 1,000 h  
NI 6115/6120 User Manual  
A-14  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Appendix A  
Specifications  
Digital I/O  
Number of channels ............................... 8 input/output  
Compatibility ......................................... TTL/CMOS  
Table A-7. Digital logic levels  
Level  
Input low voltage  
Min  
0.0 V  
2.0 V  
Max  
0.8 V  
5.0 V  
–320 µA  
10 µA  
0.4 V  
Input high voltage  
Input low current (Vin = 0 V)  
Input high current (Vin = 5 V)  
Output low voltage (IOL = 24 mA)  
Output high voltage (IOH = 13 mA)  
4.35 V  
Power-on state........................................ Input (high-impedance)  
Data transfers ......................................... DMA, interrupts,  
programmed I/O  
Input buffer ............................................ 2,000 bytes  
Output buffer.......................................... 2,000 bytes  
Transfer rate (1 word = 8 bits) ............... 10 Mwords/s  
Timing I/O  
Number of channels ............................... 2 up/down counter/timers,  
1 frequency scaler  
Resolution  
Counter/timers ................................ 24 bits  
Frequency scaler ............................. 4 bits  
Compatibility ......................................... TTL/CMOS  
Base clocks available  
Counter/timers ................................ 20 MHz, 100 kHz  
Frequency scaler ............................. 10 MHz, 100 kHz  
© National Instruments Corporation  
A-15  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Appendix A  
Specifications  
Base clock accuracy................................ 0.01%  
Max source frequency.............................20 MHz  
Min source pulse duration .....................10 ns, edge-detect mode  
Min gate pulse duration .........................10 ns, edge-detect mode  
Data transfers..........................................DMA, interrupts,  
programmed I/O  
DMA modes ...........................................Scatter-gather  
Triggers  
Analog Trigger  
NI 6115/6120 source ..............................All analog input channels,  
external trigger (PFI0/TRIG1)  
Level ....................................................... full-scale, internal;  
10 V, external  
Slope .......................................................Positive or negative  
(software selectable)  
Resolution  
NI 6115............................................8 bits, 1 in 256  
NI 6120............................................12 bits, 1 in 4,096  
Hysteresis................................................Programmable  
Bandwidth ..............................................(–3 dB) 5 MHz internal/external  
External input (PFI0/TRIG1)  
Impedance........................................10 kΩ  
Coupling ..........................................AC/DC  
Protection.........................................–0.5 V to (VCC + 0.5) V when  
configured as a digital signal,  
35 V when configured as an  
analog trigger signal or disabled,  
35 V powered off  
NI 6115/6120 User Manual  
A-16  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix A  
Specifications  
Digital Trigger  
Compatibility ......................................... TTL  
Response ................................................ Rising or falling edge  
Pulse width............................................. 10 ns min  
RTSI  
Trigger lines ........................................... 71  
Bus Interface  
Type ....................................................... Master, slave  
Power Requirement  
+5 VDC ( 5%)  
NI 6115 ........................................... 2.2 A  
NI 6120 ........................................... 3.0 A  
+3.3 V..................................................... 0.8 A  
Power available at I/O connector........... +4.65 to +5.25 VDC at 1 A  
Dimensions (not including connectors)  
Physical  
NI PCI-6115/6120........................... 31.2 by 10.6 cm  
(12.3 by 4.2 in.)  
NI PXI-6115/6120 .......................... 16 by 10 cm  
(6.3 by 3.9 in.)  
I/O connector.......................................... 68-pin male SCSI-II type  
Environmental  
Operating temperature............................ 0 to 50 °C  
Storage temperature ............................... –20 to 70 °C  
1
RTSI Trigger<6> is configured as the PXI Star Trigger for the NI PXI-6115/6120. Refer to the RTSI Triggers section of  
Chapter 3, Hardware Overview, for more information.  
© National Instruments Corporation  
A-17  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Appendix A  
Specifications  
Relative humidity ...................................10 to 90% noncondensing  
Pollution Degree (indoor use only) ........2  
Safety  
The NI 6115/6120 was evaluated using the criteria of EN 61010-1 a-2:1995  
and meets the requirements of the following standards for safety and  
electrical equipment for measurement, control and laboratory use:  
EN 61010-1, IEC 61010-1  
UL 3111-1  
CAN/CSA C22.2 No. 1010.1  
Maximum Working Voltage  
Maximum working voltage refers to the signal voltage plus the  
common-mode voltage.  
Channel-to-earth .....................................42 V, Installation Category I  
Channel-to-channel.................................42 V, Installation Category I  
Electromagnetic Compatibility  
EMC/EMI ...............................................CE, C-Tick, and FCC Part 15  
(Class A) Compliant  
Emissions................................................EN 55011 Class A at 10 m  
FCC Part 15A above 1 GHz  
Immunity ................................................Evaluated to EN 61326:1998,  
Table 1  
Note For full EMC and EMI compliance, you must operate this device with shielded  
cabling. In addition, all covers and filler panels must be installed. Refer to the Declaration  
of Conformity (DoC) for this product for any additional regulatory compliance  
information. To obtain the DoC for this product, click Declaration of Conformity at  
ni.com/hardref.nsf/. This Web site lists the DoCs by product family. Select the  
appropriate product family, followed by the product, and a link to the DoC (in Adobe  
Acrobat format) appears. Click the Acrobat icon to download or read the DoC.  
NI 6115/6120 User Manual  
A-18  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
B
Common Questions  
This appendix contains a list of commonly asked questions and their  
answers relating to usage and special features of the NI 6115/6120.  
General Information  
What is the NI 6115/6120?  
The NI 6115/6120 is a switchless and jumperless enhanced MIO device  
that uses the DAQ-STC for timing.  
What is the DAQ-STC?  
The DAQ-STC is the system timing control application-specific integrated  
circuit (ASIC) designed by NI and is the backbone of the NI 6115/6120  
device. The DAQ-STC contains seven 24-bit counters and three 16-bit  
counters. The counters are divided into the following three groups:  
AI—two 24-bit, two 16-bit counters  
AO—three 24-bit, one 16-bit counters  
General-purpose counter/timer functions—two 24-bit counters  
The groups can be configured independently with timing resolutions of  
50 ns or 10 µs. With the DAQ-STC, you can interconnect a wide variety of  
internal timing signals to other internal blocks. The interconnection scheme  
is quite flexible and completely software configurable. New capabilities  
such as buffered pulse generation and equivalent time sampling are  
possible.  
What does the maximum sampling rate mean to me?  
Maximum sampling rate is the fastest you can acquire data on the device  
and still achieve accurate results. The NI 6115 device has a maximum  
sampling rate of 10 MS/s. This sampling rate is at 10 MS/s regardless if  
1 or 4 channels are acquiring data. The NI 6120 has a maximum sampling  
rate of 800 kS/s.  
© National Instruments Corporation  
B-1  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Appendix B  
Common Questions  
What type of 5 V protection does the NI 6115/6120 have?  
The NI 6115/6120 has 5 V lines equipped with a self-resetting 1 A fuse.  
How do I use the NI 6115/6120 with the NI-DAQ C API?  
The NI-DAQ User Manual for PC Compatibles describes the general  
programming flow when using the NI-DAQ C API as well as contains  
example code. For a list of functions that support the NI 6115/6120, you  
can refer to the NI-DAQ Function Reference Help. You can access this help  
file by clicking Start»Programs»National Instruments»NI-DAQ»  
NI-DAQ Help.  
Installing and Configuring the Device  
How do you set the base address for the NI 6115/6120?  
The base address of the NI 6115/6120 is assigned automatically through the  
PCI bus protocol. This assignment is completely transparent to you.  
What jumpers should I be aware of when configuring the  
NI 6115/6120?  
The NI 6115/6120 is jumperless and switchless.  
Which NI document should I read first to get started using DAQ  
software?  
The DAQ Quick Start Guide and the NI-DAQ or application software  
release notes documentation are good places to start.  
What is the best way to test the NI 6115/6120 without programming the  
device?  
Measurement & Automation Explorer (MAX) has a Test Panel option that  
is available by selecting Devices and Interfaces and then selecting the  
device. The test panels are excellent tools for performing simple functional  
tests of the device, such as AI, DIO and counter/timer tests.  
NI 6115/6120 User Manual  
B-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Appendix B  
Common Questions  
Analog Input and Output  
Why is there a minimum sampling rate on the NI 6115?  
The NI 6115 makes use of a pipelined ADC in order to achieve high  
sampling rates. Sampling at rates below 20 kS/s can result in improper  
digitization, which appear as noise in the acquired data.  
How do I enable the programmable antialiasing filter on the  
NI 6115/6120?  
In LabVIEW, select Data Acquisition»Analog Input»Advanced Analog  
Input»AI Parameter.vi from the function palette to set the filter values of  
50 kHz and 500 kHz for the NI 6115, or to enable the 100 kHz filter for the  
NI 6120, on a per channel basis. To disable the filter, set the filter value to 0.  
Figure B-1. Setting Filter Values in LabVIEW  
In NI-DAQ, use the AI_Change_Parameter function to set the filter  
value. Set paramID to ND_Digital_Filter. Set ParamValue to  
ND_High for a filter value of 500 kHz on the NI 6115 or 100 kHz on the  
NI 6120. Use ND_Low for a filter value of 50 kHz on the NI 6115. Use  
ND_None to disable the filter. The filter is disabled by default.  
I have connected a differential input signal, but my readings are  
random and drift rapidly. What is wrong?  
Check your ground-reference connections. The signal may be referenced  
to a level that is considered floating with reference to the device ground  
reference. Even if you are in differential mode, the signal must still be  
referenced to the same ground level as the device reference. There are  
various methods of achieving this reference while maintaining a high  
common-mode rejection ratio (CMRR). These methods are outlined in  
Chapter 4, Connecting Signals.  
© National Instruments Corporation  
B-3  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Appendix B  
Common Questions  
I’m using the DACs to generate a waveform, but I discovered with a  
digital oscilloscope that there are glitches on the output signal. Is this  
normal?  
When the DAC switches from one voltage to another, any DAC produces  
glitches due to released charges. The largest glitches occur when the most  
significant bit (MSB) of the D/A code switches. You can build a lowpass  
deglitching filter to remove some of these glitches, depending on the  
frequency and nature of your output signal.  
Can I synchronize a one-channel AI data acquisition with a  
one-channel AO waveform generation on the NI 6115/6120?  
Yes. One way to accomplish synchronization is to use the waveform  
generation timing pulses to control the AI data acquisition. To do this,  
follow steps 1 through 4 below, in addition to the usual steps for data  
acquisition and waveform generation configuration.  
1. Enable the PFI5 line for output, as follows:  
If you are using NI-DAQ, call  
Select_Signal(deviceNumber, ND_PFI_5,  
ND_OUT_UPDATE, ND_HIGH_TO_LOW).  
If you are using LabVIEW, select Data Acquisition»Calibration  
and Configuration»Route Signal.vi from the function palette  
and set signal name to PFI5and signal source to AO Update.  
2. Set up data acquisition timing so that the timing signal for A/D  
conversion comes from PFI5, as follows:  
If you are using NI-DAQ, call  
Select_Signal(deviceNumber, ND_IN_SCAN_START,  
ND_PFI_7, ND_HIGH_TO_LOW).  
If you are using LabVIEW, select Data Acquisition»Analog  
Input»Advanced Analog Input»AI Clock Config.vi with clock  
source code set to PFI pin, high to low, and clock source  
string set to 5.  
3. Initiate AI data acquisition, which starts only when the AO waveform  
generation starts.  
4. Initiate AO waveform generation.  
NI 6115/6120 User Manual  
B-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Appendix B  
Common Questions  
Timing and Digital I/O  
What types of triggering can be hardware-implemented on the  
NI 6115/6120?  
Hardware digital and analog triggering are both supported on the  
NI 6115/6120.  
If I’m using one of the general-purpose counter/timers on the  
NI 6115/6120, but I do not see the counter/timer output on the I/O  
connector, what am I doing wrong?  
If you are using NI-DAQ or LabWindows/CVI, you must configure the  
output line to output the signal to the I/O connector. Use the  
Select_Signalcall in NI-DAQ to configure the output line. By default,  
all timing I/O lines except EXTSTROBE* are high-impedance.  
What are the PFIs and how do I configure these lines?  
PFIs are Programmable Function Inputs. These lines serve as connections  
to virtually all internal timing signals.  
If you are using NI-DAQ or LabWindows/CVI, use the Select_Signal  
function to route internal signals to the I/O connector, route external signals  
to internal timing sources, or tie internal timing signals together.  
If you are using NI-DAQ with LabVIEW and you want to connect external  
signal sources to the PFI lines, you can use AI Clock Config, AI Trigger  
Config, AO Clock Config, AO Trigger and Gate Config, and Counter Set  
Attribute advanced-level VIs to indicate which function the connected  
signal serves. Use the Route Signal VI to enable the PFI lines to output  
internal signals.  
Table B-1. Signal Name Equivalencies  
Hardware  
LabVIEW  
Signal Name  
Route Signal  
AI Start Trigger  
AI Stop Trigger  
AI Scan Start  
NI-DAQ Select_Signal  
ND_IN_START_TRIGGER  
TRIG1  
TRIG2  
ND_IN_STOP_TRIGGER  
ND_IN_SCAN_START  
STARTSCAN  
SISOURCE  
CONVERT*  
ND_IN_SCAN_CLOCK_TIMEBASE  
ND_IN_CONVERT  
AI Convert  
© National Instruments Corporation  
B-5  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Appendix B  
Common Questions  
Table B-1. Signal Name Equivalencies (Continued)  
Hardware  
Signal Name  
LabVIEW  
Route Signal  
NI-DAQ Select_Signal  
AIGATE  
AO Start Trigger  
AO Update  
ND_IN_EXTERNAL_GATE  
ND_OUT_START_TRIGGER  
ND_OUT_UPDATE  
WFTRIG  
UPDATE*  
UISOURCE  
AOGATE  
ND_OUT_UPDATE_CLOCK_TIMEBASE  
ND_OUT_EXTERNAL_GATE  
Caution If you enable a PFI line for output, do not connect any external signal source to it;  
if you do, you can damage the device, the computer, and the connected equipment.  
What are the power-on states of the PFI and DIO lines on the  
I/O connector?  
At system power-on and reset, both the PFI and DIO lines are set to  
high-impedance by the hardware. Hence, the device circuitry is not actively  
driving the output either high or low. However, these lines may have pull-up  
or pull-down resistors connected to them as shown in Table 4-5, Digital I/O  
Signal Summary. These resistors weakly pull the output to either a  
logic-high or logic-low state. For example, DIO(0) is in the  
high-impedance state after power on, and Table 4-5 shows that there is a  
50 kpull-up resistor. This pull-up resistor sets the DIO(0) pin to a logic  
high when the output is in a high-impedance state.  
NI 6115/6120 User Manual  
B-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
C
Technical Support and  
Professional Services  
Visit the following sections of the NI Web site at ni.comfor technical  
support and professional services:  
Support—Online technical support resources include the following:  
Self-Help Resources—For immediate answers and solutions,  
visit our extensive library of technical support resources available  
in English, Japanese, and Spanish at ni.com/support. These  
resources are available for most products at no cost to registered  
users and include software drivers and updates, a KnowledgeBase,  
product manuals, step-by-step troubleshooting wizards, hardware  
schematics and conformity documentation, example code,  
tutorials and application notes, instrument drivers, discussion  
forums, a measurement glossary, and so on.  
Assisted Support Options—Contact NI engineers and other  
measurement and automation professionals by visiting  
ni.com/ask. Our online system helps you define your question  
and connects you to the experts by phone, discussion forum,  
or email.  
Training—Visit ni.com/custedfor self-paced tutorials, videos, and  
interactive CDs. You also can register for instructor-led, hands-on  
courses at locations around the world.  
System Integration—If you have time constraints, limited in-house  
technical resources, or other project challenges, NI Alliance Program  
members can help. To learn more, call your local NI office or visit  
ni.com/alliance.  
Declaration of Conformity (DoC)—A DoC is our claim of  
compliance with various European Council Directives using the  
manufacturer’s self-declaration of conformance. This system affords  
the user protection for electronic compatibility (EMC) and product  
safety. You can obtain the DoC for your product by visiting  
ni.com/hardref.nsf.  
© National Instruments Corporation  
C-1  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                         
Appendix C  
Technical Support and Professional Services  
Calibration Certificate—If your product supports calibration, you  
can obtain the calibration certificate for your product at  
ni.com/calibration.  
If you searched ni.comand could not find the answers you need, contact  
your local office or NI corporate headquarters. Phone numbers for our  
worldwide offices are listed at the front of this manual. You also can visit  
the Worldwide Offices section of ni.com/niglobalto access the branch  
office Web sites, which provide up-to-date contact information, support  
phone numbers, email addresses, and current events.  
NI 6115/6120 User Manual  
C-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Glossary  
Prefix  
p-  
Meaning  
pico-  
Value  
10–12  
10–9  
10– 6  
10–3  
103  
n-  
nano-  
micro-  
milli-  
µ-  
m-  
k-  
kilo-  
M-  
mega-  
106  
Numbers/Symbols  
°
degree  
>
<
/
greater than  
greater than or equal to  
less than  
less than or equal to  
per  
%
percent  
plus or minus  
positive of, or plus  
negative of, or minus  
ohm  
+
square root of  
+5 VDC source signal  
+5 V  
© National Instruments Corporation  
G-1  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Glossary  
A
A
amperes  
A/D  
analog-to-digital  
AC  
alternating current  
ACH  
ACH0GND  
ADC  
analog input channel signal  
analog input channel ground signal  
analog-to-digital converter—an electronic device, often an integrated  
circuit, that converts an analog voltage to a digital number  
ADE  
application development environment such as LabVIEW,  
LabWindows/CVI, Measurement Studio, Visual Basic, C, and C++  
AI  
analog input  
AIGATE  
aliasing  
analog input gate signal  
the consequence of sampling that causes signals with frequencies higher  
than half the sampling frequency to appear as lower frequency  
components in a frequency spectrum  
ANSI  
AO  
American National Standards Institute  
analog output  
AOGND  
ASIC  
analog output ground signal  
application-specific integrated circuit—a proprietary semiconductor  
component designed and manufactured to perform a set of specific  
functions  
B
Bessel filter  
a filter with a maximally flat response in both magnitude and phase. The  
phase response in the passband, which is usually the region of interest, is  
nearly linear. Bessel filters reduce nonlinear phase distortion inherent in  
all IIR filters.  
bipolar  
a signal range that includes both positive and negative values  
NI 6115/6120 User Manual  
G-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
C
C
Celsius  
CalDAC  
cm  
calibration DAC  
centimeter  
CMOS  
CMRR  
complementary metal-oxide semiconductor  
common-mode rejection ratio—a measure of an instrument to reject  
interference from a common-mode signal, usually expressed in  
decibels (dB)  
CompactPCI  
CONVERT*  
correlated DIO  
counter/timer  
CTR  
a Eurocard configuration of the PCI bus for industrial applications  
convert signal  
can clock digital I/O on the same clock as analog I/O  
a circuit that counts external pulses or clock pulses (timing)  
counter  
D
D/A  
digital-to-analog  
DAC  
digital-to-analog converter—an electronic device that converts a digital  
number into a corresponding analog voltage or current  
DAC0OUT  
DAC1OUT  
DAQ  
analog channel 0 output signal  
analog channel 1 output signal  
data acquisition—(1) collecting and measuring electrical signals from  
sensors, transducers, and test probes or fixtures and inputting them to a  
computer for processing; (2) collecting and measuring the same kinds of  
electrical signals with A/D and/or DIO devices plugged into a computer,  
and possibly generating control signals with D/A and/or DIO devices in  
the same computer  
© National Instruments Corporation  
G-3  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
DAQ-STC  
data acquisition system timing controller—an application-specific  
integrated circuit (ASIC) for the system timing requirements of a general  
A/D and D/A system, such as a system containing the National  
Instruments E Series devices  
dB  
decibel—the unit for expressing a logarithmic measure of the ratio of  
two signal levels: dB = 20log10 V1/V2, for signals in volts  
DC  
direct current  
DGND  
DI  
digital ground signal  
digital input  
DIFF  
differential mode—an analog input mode consisting of two terminals,  
both of which are isolated from computer ground, whose difference is  
measured  
DIO  
digital input/output  
dual inline package  
DIP  
dithering  
the addition of Gaussian noise to an analog input signal for the purpose  
of increasing the resolution of a measurement when using averaging  
DMA  
direct memory access—a method by which data can be transferred  
to/from computer memory from/to a device or memory on the bus while  
the processor does something else; the fastest method of transferring data  
to/from computer memory  
DNL  
DO  
differential nonlinearity—a measure in least significant bit of the  
worst-case deviation of code widths from their ideal value of 1 LSB  
digital output  
E
EEPROM  
electrically erasable programmable read-only memory—ROM that can  
be erased with an electrical signal and reprogrammed  
ENOB  
effective number of bits  
NI 6115/6120 User Manual  
G-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
ESD  
electrostatic discharge  
external strobe signal  
EXTSTROBE*  
F
F
farad—a measurement unit of capacitance  
FIFO  
first-in first-out memory buffer—the first data stored is the first data sent  
to the acceptor; often used on DAQ devices to temporarily store incoming  
or outgoing data until that data can be retrieved or output  
floating signal sources  
signal sources, including batteries, transformers, or thermocouples, with  
voltage signals that are not connected to an absolute reference or system  
ground; also called nonreferenced signal sources  
FPGA  
field programmable gate array  
frequency output signal  
FREQ_OUT  
G
gain  
the factor by which a signal is amplified, sometimes expressed in decibels  
gate signal  
GATE  
GPCTR  
general-purpose counter signal  
GPCTR0_GATE  
GPCTR0_OUT  
GPCTR0_SOURCE  
GPCTR0_UP_DOWN  
GPCTR1_GATE  
GPCTR1_OUT  
GPCTR1_SOURCE  
general-purpose counter 0 gate signal  
general-purpose counter 0 output signal  
general-purpose counter 0 clock source signal  
general-purpose counter 0 up down signal  
general-purpose counter 1 gate signal  
general-purpose counter 1 output signal  
general-purpose counter 1 clock source signal  
© National Instruments Corporation  
G-5  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
GPCTR1_UP_DOWN  
general-purpose counter 1 up down signal  
grounded signal  
sources  
signal sources with voltage signals that are referenced to a system ground,  
such as the earth or a building ground; also called grounded signal  
sources  
H
h
hour  
Hz  
hertz—the number of scans read or updates written per second  
I
I/O  
input/output—the transfer of data to/from a computer system involving  
communications channels, operator interface devices, and/or data  
acquisition and control interfaces  
impedance  
in.  
resistance  
inch or inches  
INL  
integral nonlinearity—a measurement in least significant bits of the  
worst-case deviation from the ideal A/D or D/A transfer characteristic of  
the analog I/O circuitry  
IOH  
current, output high  
current, output low  
interrupt request  
IOL  
IRQ  
K
kHz  
kilohertz  
NI 6115/6120 User Manual  
G-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
L
LabVIEW  
Laboratory Virtual Instrument Engineering Workbench—a program  
development application based on the programming language G and used  
commonly for test and measurement purposes  
LED  
LSB  
light emitting diode  
least significant bit  
M
master  
a functional part of a MXI/VME/VXIbus device that initiates data  
transfers on the backplane; a transfer can be either a read or a write  
MAX  
Measurement and Automation Explorer  
megabytes of memory  
MB  
Measurement Studio  
a set of test and measurement-oriented software tools from National  
Instruments for C, C++, and Visual Basic users  
MHz  
MIO  
MITE  
MSB  
mux  
megahertz  
multifunction I/O  
MXI Interface to Everything  
most significant bit  
multiplexer—a switching device with multiple inputs that sequentially  
connects each of its inputs to its output, typically at high speeds, in order  
to measure several signals with a single analog input channel  
mV  
millivolts  
MXI  
a high-performance communication link that interconnects devices using  
round, flexible cables  
© National Instruments Corporation  
G-7  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
N
NC  
not connected (signal)  
NI  
National Instruments  
NI-DAQ  
noise  
National Instruments driver software for DAQ hardware  
an undesirable electrical signal from external sources such as the AC  
power line, motors, generators, transformers, fluorescent lights, soldering  
irons, CRT displays, computers, electrical storms, welders, radio  
transmitters, and internal sources such as semiconductors, resistors, and  
capacitors; noise corrupts signals you are trying to send or receive  
nonreferenced signal  
sources  
signal sources, including batteries, transformers, or thermocouples, with  
voltage signals that are not connected to an absolute reference or system  
ground; also called nonreferenced signal sources  
Nyquist frequency  
the maximum signal frequency that a sampling system can accurately  
represent in frequency spectrum measurement, which is half the sampling  
frequency  
O
OUT  
output pin—a counter output pin where the counter can generate various  
TTL pulse waveforms  
P
PCI  
Peripheral Component Interconnect—a high-performance expansion  
bus architecture originally developed by Intel to replace ISA and EISA;  
is achieving widespread acceptance as a standard for PCs and  
work-stations, and offers a theoretical maximum transfer rate of  
132 Mbytes/s  
pd  
pull down  
PFI  
programmable function input  
PFI0/trigger 1  
PFI0/TRIG1  
PFI1/TRIG2  
PFI1/trigger 2  
NI 6115/6120 User Manual  
G-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
PFI2/CONVERT*  
PFI2/convert  
PFI3/GPCTR1_SOURCE PFI3/general purpose counter 1 source  
PFI4/GPCTR1_GATE  
PFI5/UPDATE*  
PFI4/general-purpose counter 1 gate  
PFI5/update  
PFI6/WFTRIG  
PFI6/waveform trigger  
PFI7/start of scan  
PFI7/STARTSCAN  
PFI8/GPCTR0_SOURCE PFI8/general-purpose counter 0 source  
PFI9/GPCTR0_GATE  
PFI9/general-purpose counter 0 gate  
programmable gain instrumentation amplifier  
phase-locked loop  
PGIA  
PLL  
ppm  
parts per million  
pseudodifferential  
channels  
pseudodifferential channels are all referred to a common ground, but this  
ground is not directly connected to the computer ground. Often this  
connection is made by a relatively low value resistor to give some  
isolation between the two grounds.  
pu  
pull up  
PWB  
PXI  
printed wire board  
PCI eXtensions for Instrumentation—an open specification that builds  
off the CompactPCI specification by adding instrumentation-specific  
features  
R
range  
the maximum and minimum parameters between which a sensor,  
instrument, or device operates with a specified set of characteristics  
referenced signal  
sources  
signal sources with voltage signals that are referenced to a system ground,  
such as the earth or a building ground; also called grounded signal  
sources  
© National Instruments Corporation  
G-9  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
rise time  
the difference in time between the 10% and 90% points of the step  
response of a system  
rms  
root mean square  
RTD  
resistive temperature detector—a metallic probe that measures  
temperature based upon its coefficient of resistivity  
RTSI bus  
real-time system integration bus—the National Instruments timing bus  
that connects DAQ devices directly, by means of connectors on top of the  
devices for precise synchronization of functions  
RTSI_OSC  
RTSI Oscillator—RTSI bus master clock  
S
s
seconds  
samples  
S
S/s  
samples per second—used to express the rate at which a DAQ device  
samples an analog signal  
SCANCLK  
scan clock signal  
scatter-gather  
a term that describes very high-speed DMA burst-mode transfers that are  
made only by the bus master  
signal conditioning  
SFDR  
the manipulation of signals to prepare them for digitizing  
spurious free dynamic range  
SI counter clock signal  
SISOURCE  
SOURCE  
source signal  
STARTSCAN  
STC  
start scan signal  
system timing controller  
system noise  
a measure of the amount of noise present in an analog circuit or when the  
NI 6115/6120 User Manual  
G-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
T
TC  
terminal count—the ending value of a counter  
tgh  
gate hold time  
tgsu  
tgw  
THD  
gate setup time  
gate pulse width  
total harmonic distortion—the ratio of the total rms signal due to  
harmonic distortion to the overall rms signal, in decibel or a percentage  
thermocouple  
toff  
a temperature sensor created by joining two dissimilar metals whose  
junction produces a small voltage as a function of the temperature  
an offset (delayed) pulse; the offset is t nanoseconds from the falling edge  
of the CONVERT* signal  
tout  
tp  
output delay time  
period of a pulse train  
trigger signal  
TRIG  
tsc  
source clock period  
source pulse width  
transistor-transistor logic  
pulse width  
tsp  
TTL  
tw  
U
UI  
update interval  
UISOURCE  
UPDATE*  
update interval counter clock signal  
update signal  
© National Instruments Corporation  
G-11  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
V
V
volts  
VCC  
Vcm  
VCXO  
VDC  
VI  
collector common voltage—power supply voltage  
common-mode noise and ground potential  
voltage-controlled crystal oscillator  
volts direct current  
virtual instrument—(1) a combination of hardware and/or software  
elements, typically used with a PC, that has the functionality of a classic  
stand-alone instrument; (2) a LabVIEW software module (VI), which  
consists of a front panel user interface and a block diagram program  
Vin  
volts in  
Vm  
measured voltage  
volts, output high  
volts, output low  
volts out  
VOH  
VOL  
VOUT  
Vrms  
Vs  
volts, root mean square  
ground-referenced signal source  
W
WFTRIG  
waveform generation trigger signal  
NI 6115/6120 User Manual  
G-12  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
analog input specifications  
DC transfer characteristics, A-2  
dynamic characteristics, A-4  
stability, A-11  
Numerics  
+5 V signal  
self-resetting fuse, 4-18, B-2  
analog output  
overview, 3-5  
questions about, B-3  
A
analog output specifications  
DC transfer characteristics, A-12  
stability, A-14  
above-high-level analog triggering mode, 3-7  
AC input coupling, 3-4  
ACH<0..3>– signal  
analog input signal connections, 4-8  
description (table), 4-3  
signal summary (table), 4-5  
ACH<0..3>+ signal  
analog input signal connections, 4-8  
description (table), 4-3  
above-high-level analog triggering mode  
below-low-level analog triggering mode  
block diagram, 3-6  
(figure), 3-7  
(figure), 3-7  
(figure), 3-8  
signal summary (table), 4-5  
ACH<0..3>GND signal (table), 4-3  
AIGATE signal  
gating DAQ sequences, 4-25, 4-27  
overview, 4-27  
RTSI bus signal connections (figure), 3-13  
analog input  
input coupling, 3-4  
input mode, 3-2  
input polarity and range, 3-3  
questions about, B-3  
selection considerations, 3-4  
analog input signal connections  
common-mode signal rejection, 4-12  
floating signal sources, 4-11  
ground-referenced signal sources, 4-9  
nonreferenced signal sources, 4-11  
pseudodifferential connections  
definition, 4-8  
antialiasing filters  
description, 3-8  
effects of hardware and software filtering  
(figure), 3-9  
enabling, B-3  
ground referenced signals (figure), 4-9  
© National Instruments Corporation  
I-1  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Index  
AOGND signal  
C
See also I/O connectors  
custom cabling, 1-6  
field wiring considerations, 4-39  
optional equipment, 1-6  
calibration  
B
base address for NI 6115/6120 device, B-2  
below-low-level analog triggering mode, 3-6  
bipolar input, 3-3  
external calibration, 5-2  
loading calibration constants, 5-1  
self-calibration, 5-2  
block diagrams  
correlating DIO signals, 4-16  
common-mode signal rejection, 4-12  
configuration  
description, 2-3  
questions about, B-2  
connectors. See I/O connectors  
contacting National Instruments, C-2  
analog trigger, 3-6  
NI 6115 block diagram, 3-1  
NI 6120 block diagram, 3-2  
phase-locked loop circuit (figure), 3-10  
bus  
CompactPCI  
master device slot support (note), 2-2  
using PXI with CompactPCI, 1-2  
interface specifications, A-17  
PCI  
overview, 1-1  
PXI, 1-1  
assignments (table), 1-3  
See also PFI2/CONVERT* signal  
output timing (figure), 4-26  
(figure), 3-13  
(figure), 4-21  
RTSI  
overview, 1-1  
PCI RTSI Bus Signal Connection  
(figure), 3-13  
(figure), 4-21  
RTSI triggers, 3-12  
timing signal routing, 3-10  
using PXI with CompactPCI, 1-2  
correlated digital I/O. See digital I/O  
counter/timer applications, B-5  
custom cabling, 1-6  
customer  
education, C-1  
NI 6115/6120 User Manual  
I-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
digital I/O  
correlated  
D
DAC0OUT signal  
analog output signal connections, 4-14  
description (table), 4-3  
signal summary (table), 4-5  
DAC1OUT signal  
DAQ timing connections. See data acquisition  
timing connections  
rising-edge AO update signal driving  
DIO signal (figure), 4-17  
signals for clock source, 4-16  
questions about, B-5  
signal connections, 4-15  
specifications, A-15  
DAQ-STC system timing controller  
overview, 1-1  
digital trigger  
overview, 3-8  
questions about, B-1  
specifications, A-17  
timing signal routing, 3-10  
data acquisition timing connections  
AIGATE signal, 4-20, 4-27  
CONVERT* signal, 4-20, 4-26  
EXTSTROBE* signal, 4-29  
SISOURCE signal, 4-20, 4-27  
DIO<0..7> signal  
digital I/O signal connections, 4-15  
signal summary (table), 4-6  
documentation  
about this manual, xi  
conventions used in manual, xi  
National Instruments documentation, xii  
online library, C-1  
typical pretriggered acquisition  
(figure), 4-21  
related documentation, xiii  
instrument, C-1  
software, C-1  
DC input coupling, 3-4  
Declaration of Conformity, C-1  
deglitching, questions about, B-4  
device clocks, 3-12  
electromagnetic compatibility  
specifications, A-18  
environmental noise, avoiding, 4-39  
environmental specifications, A-17  
equipment, optional, 1-6  
device configuration. See configuration  
DGND signal  
description (table), 4-3  
digital I/O signal connections, 4-15  
power connections, 4-18  
example code, C-1  
© National Instruments Corporation  
I-3  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
external calibration, 5-2  
EXTSTROBE* signal  
signal summary (table), 4-6  
(figure), 3-13  
general-purpose counter timing summary  
(figure), 4-37  
general-purpose timing connections, 4-34  
F
field wiring considerations, 4-39  
floating signal sources  
signal summary (table), 4-7  
See also PFI8/GPCTR0_SOURCE signal  
(figure), 4-37  
description, 4-7  
signal connections, 4-11  
description (table), 4-5  
(figure), 4-34  
RTSI bus signal connections  
(figure), 3-13  
general-purpose timing connections, 4-38  
signal summary (table), 4-7  
frequently asked questions. See questions and  
answers  
G
general-purpose timing connections, 4-35  
general-purpose timing signal connections  
FREQ_OUT signal, 4-38  
See also PFI4/GPCTR1_GATE signal  
GPCTR0_GATE signal, 4-33  
GPCTR0_OUT signal, 4-34  
GPCTR1_OUT signal  
GPCTR0_SOURCE signal, 4-32  
GPCTR0_UP_DOWN signal, 4-35  
GPCTR1_GATE signal, 4-36  
GPCTR1_OUT signal, 4-36  
GPCTR1_SOURCE signal, 4-35  
GPCTR1_UP_DOWN signal, 4-37  
description (table), 4-4  
(figure), 4-37  
GPCTR1_SOURCE signal  
general purpose timing connections, 4-35  
general-purpose counter timing summary  
(figure), 4-37  
glitches, questions about, B-4  
GPCTR0_GATE signal  
See also PFI9/GPCTR0_GATE signal  
(figure), 4-37  
general-purpose timing connections, 4-35  
GPCTR1_OUT signal timing  
(figure), 4-37  
general-purpose timing connections, 4-33  
NI 6115/6120 User Manual  
I-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
GPCTR1_UP_DOWN signal  
digital I/O lines, 3-10  
general-purpose timing connections, 4-37  
ground-referenced signal sources  
description, 4-8  
I/O connectors  
exceeding maximum ratings  
(caution), 4-1  
overview, 4-1  
input coupling, 3-4  
input polarity and range  
bipolar input, 3-3  
H
hardware installation  
procedure, 2-1  
(table), 3-4  
overvoltage hazard (caution), 3-3  
selection considerations, 3-4  
inside-region analog triggering mode, 3-7  
installation  
unpacking NI 6115/6120, 1-8  
hardware overview  
analog input  
input coupling, 3-4  
input mode, 3-2  
input polarity and range, 3-3  
selection considerations, 3-4  
analog output, 3-5  
category, 1-9  
questions about, B-2  
software installation, 2-1  
unpacking NI 6115/6120, 1-7  
internal timebase, device and RTSI  
clocks, 3-12  
analog trigger  
block diagram, 3-6  
overview, 3-5  
antialiasing filters, 3-8  
block diagrams  
NI 6115 block diagram, 3-1  
NI 6120 block diagram, 3-2  
correlated digital I/O, 3-10  
phase-locked loop circuit, 3-9  
clocks, 3-12  
L
loading calibration constants, 5-1  
lowValue, 3-5  
(figure), 3-11  
M
help  
manual. See documentation  
professional services, C-1  
technical support, C-1  
high-hysteresis analog triggering mode, 3-7  
highValue, 3-5  
MITE bus interface chip, 1-1  
© National Instruments Corporation  
I-5  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
N
National Instruments  
PFI0/TRIG1 signal  
calibration certificate, C-2  
customer education, C-1  
Declaration of Conformity, C-1  
documentation, xii  
professional services, C-1  
system integration services, C-1  
technical support, C-1  
See also TRIG1 signal  
analog triggering, 3-5  
description (table), 4-4  
signal summary (table), 4-6  
PFI1/TRIG2 signal  
See also TRIG2 signal  
description (table), 4-4  
worldwide offices, C-2  
NI 6115/6120 device  
signal summary (table), 4-6  
PFI2/CONVERT* signal  
See also CONVERT* signal  
description (table), 4-4  
See also hardware overview  
block diagrams  
NI 6115 block diagram, 3-1  
NI 6120 block diagram, 3-2  
configuration, 2-3  
signal summary (table), 4-6  
PFI3/GPCTR1_SOURCE signal  
See also GPCTR1_SOURCE signal  
description (table), 4-4  
custom cabling, 1-6  
optional equipment, 1-6  
overview, 1-1  
questions about  
analog input and output, B-3  
general information, B-1  
installation and configuration, B-2  
timing and digital I/O, B-5  
requirements for getting started, 1-3  
safety information, 1-8  
unpacking, 1-7  
signal summary (table), 4-7  
PFI4/GPCTR1_GATE signal  
See also GPCTR1_GATE signal  
description (table), 4-4  
signal summary (table), 4-7  
PFI5/UPDATE* signal  
See also UPDATE* signal  
description (table), 4-4  
signal summary (table), 4-7  
PFI6/WFTRIG signal  
NI-DAQ driver software  
See also WFTRIG signal  
description (table), 4-4  
noise, avoiding, 4-39  
nonreferenced signal connections, 4-11  
signal summary (table), 4-7  
PFI7/STARTSCAN signal  
See also STARTSCAN signal  
description (table), 4-4  
signal summary (table), 4-7  
PFI8/GPCTR0_SOURCE signal  
See also GPCTR0_SOURCE signal  
description (table), 4-5  
O
online technical support, C-1  
optional equipment, 1-6  
signal summary (table), 4-7  
NI 6115/6120 User Manual  
I-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
PFI9/GPCTR0_GATE signal  
See also GPCTR0_GATE signal  
description (table), 4-5  
signal summary (table), 4-7  
PFIs (programmable function inputs)  
questions about, B-5  
questions and answers  
analog input and output, B-3  
general information, B-1  
signal name equivalencies (table), B-5  
signal routing, 3-10  
timing connections, 4-20  
phase-locked loop circuit  
Real-Time System Integration. See RTSI  
related documentation, xiii  
requirements for getting started, 1-3  
RTSI  
block diagram, 3-10  
description, 3-9  
phone technical support, C-2  
physical specifications, A-17  
pin assignments  
bus signal connections (figure), 3-13  
clocks  
I/O connector (figure), 4-2  
(table), 1-3  
correlating DIO signals, 4-16  
description, 3-12  
overview, 1-1  
PLL. See phase-locked loop circuit  
polarity. See input polarity and range  
posttriggered data acquisition, 4-20  
power connections  
triggers  
specifications, A-17  
+5 V power pins, 4-18, B-2  
incorrect connections (caution), 4-18  
power-on states of PFI and DIO lines, B-6  
self-resetting fuse, 4-18, B-2  
power-on states of PFI and DIO lines, B-6  
safety information, 1-8  
sampling rate  
maximum, B-1  
minimum, B-3  
programmable function inputs (PFIs). See  
PFIs (programmable function inputs)  
definition, 4-8  
(figure), 4-21  
typical pretriggered acquisition, 4-21  
description (table), 4-3  
signal summary (table), 4-6  
timing connections, 4-28  
self-calibration, 5-2  
ground-referenced signals (figure), 4-9  
PXI  
PXI-6115/6120 J2 pin assignments  
(table), 1-3  
using with CompactPCI, 1-2  
self-resetting fuse, 4-18, B-2  
© National Instruments Corporation  
I-7  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
signal connections  
overview, 4-1  
power connections, 4-18  
connections, 4-20  
common-mode signal rejection, 4-12  
sources, 4-9  
nonreferenced signal sources, 4-11  
analog output connections, 4-14  
data acquisition timing connections  
AIGATE signal, 4-27  
data acquisition timing  
connections, 4-32  
connections, 4-29  
CONVERT* signal, 4-26  
EXTSTROBE* signal, 4-29  
SISOURCE signal, 4-27  
TRIG1 signal, 4-21  
(figure), 4-21  
types of signal sources  
floating, 4-7  
ground-referenced, 4-8  
waveform generation timing connections  
working voltage range, 4-13  
SISOURCE signal, 4-27  
digital I/O  
connections, 4-15  
software drivers, C-1  
software installation, 2-1  
(table), 3-4  
connections  
overview, 3-3  
specifications  
FREQ_OUT signal, 4-38  
GPCTR0_GATE signal, 4-33  
GPCTR0_OUT signal, 4-34  
GPCTR0_SOURCE signal, 4-32  
GPCTR0_UP_DOWN signal, 4-35  
GPCTR1_OUT signal, 4-36  
GPCTR1_SOURCE signal, 4-35  
I/O connectors  
analog input  
DC transfer characteristics, A-2  
dynamic characteristics, A-4  
input characteristics, A-1  
stability, A-11  
analog output  
DC transfer characteristics, A-12  
dynamic characteristics, A-14  
output characteristics, A-12  
stability, A-14  
exceeding maximum ratings  
(caution), 4-1  
voltage output, A-13  
NI 6115/6120 User Manual  
I-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
bus interface, A-17  
digital I/O, A-15  
electromagnetic compatibility, A-18  
environmental, A-17  
physical, A-17  
EXTSTROBE* signal, 4-29  
SISOURCE signal, 4-27  
TRIG1 signal, 4-21  
power requirements, A-17  
RTSI trigger lines, A-17  
safety, A-18  
TRIG2 signal, 4-22  
timing I/O, A-15  
triggers  
connections  
analog trigger, A-16  
digital trigger, A-17  
FREQ_OUT signal, 4-38  
GPCTR0_GATE signal, 4-33  
GPCTR0_OUT signal, 4-34  
GPCTR0_SOURCE signal, 4-32  
stability specifications  
analog input, A-11  
input timing (figure), 4-24  
RTSI bus signal connections  
GPCTR1_SOURCE signal, 4-35  
connections, 4-20  
timing connections, 4-24  
typical pretriggered acquisition  
(figure), 4-21  
questions about, B-5  
timing I/O connections (figure), 4-19  
waveform generation timing connections  
UISOURCE signal, 4-31  
UPDATE* signal, 4-30  
WFTRIG signal, 4-29  
using the SISOURCE signal, 4-27  
support  
timing I/O  
questions about, B-5  
timing signal routing  
clocks, 3-12  
T
technical support, C-1  
programmable function inputs, 3-12  
RTSI triggers, 3-12  
STARTSCAN* signal routing  
(figure), 3-11  
telephone technical support, C-2  
timebase (clocks), 3-12  
timing connections  
data acquisition timing connections  
AIGATE signal, 4-27  
training  
customer, C-1  
CONVERT* signal, 4-26  
© National Instruments Corporation  
I-9  
NI 6115/6120 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
TRIG1 signal  
RTSI bus signal connections  
(figure), 3-13  
See also PFI0/TRIG1 signal  
using with UISOURCE signal, 4-31  
output timing (figure), 4-22  
(figure), 3-13  
timing connections, 4-21  
typical pretriggered acquisition  
V
VCC signal (table), 4-6  
voltage  
output specifications, A-13  
working voltage range, 4-13  
(VCXO), 3-9  
TRIG2 signal  
See also PFI1/TRIG2 signal  
output timing (figure), 4-23  
RTSI bus signal connections  
W
waveform generation timing connections  
UISOURCE signal, 4-31  
UPDATE* signal, 4-30  
WFTRIG signal, 4-29  
waveform generation, questions about, B-4  
Web  
(figure), 4-21  
triggers  
analog trigger  
description, 3-5  
professional services, C-1  
specifications, A-16  
questions about, B-5  
RTSI bus signal connections  
(figure), 3-13  
U
(figure), 3-13  
timing connections, 4-29  
wiring considerations, 4-39  
working voltage range, 4-13  
worldwide technical support, C-2  
unpacking NI 6115/6120, 1-7  
UPDATE* signal  
See also PFI5/UPDATE* signal  
input timing (figure), 4-31  
output timing (figure), 4-31  
NI 6115/6120 User Manual  
I-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Milwaukee Sander HEAVY DUTYSANDERS User Manual
Mr Coffee Coffeemaker JWX9 User Manual
Muratec All in One Printer I Fax User Manual
NEC All in One Printer 703 A0368 001 User Manual
Nikon Projector NP4100 User Manual
Nilfisk Advance America Vacuum Cleaner SDV 4500 User Manual
Omega Speaker Systems Stud Sensor CY670 Series User Manual
Orion Car Audio Universal Remote OE331 21 User Manual
Panasonic Blu ray Player DMP BDT105 User Manual
Panasonic Microwave Oven IP3999_38N04AP_01_111206qxd User Manual