NI PXI/PCI-5114 Specifications
8-Bit 250 MS/s Digitizer
This document lists the specifications for the NI PXI/PCI-5114 (NI 5114)
high-speed digitizer. Unless otherwise noted, these specifications are valid
for the following conditions:
•
•
•
All filter settings
All impedance selections
Sample clock set to 250 MS/s
Typical values are representative of an average unit operating at room
temperature. Specifications are subject to change without notice. For the
most recent NI 5114 specifications, visit ni.com/manuals.
To access the NI 5114 documentation, including the NI High-Speed
of the NI 5114 signals, navigate to Start»All Programs»National
Instruments»NI-SCOPE»Documentation.
cause burns. Allow the NI 5114 to cool before removing it from the PXI chassis or PC.
Phase-Locked Loop (PLL) Reference Clock................................... 7
Trigger..................................................................................................... 8
Reference (Stop) Trigger ................................................................. 8
TRIG (External Trigger, Front Panel Connector)............................ 10
PFI 0 and PFI 1 (Programmable Function Interface,
AUX Front Panel Connectors)............................................................. 11
TClk Specifications................................................................................. 12
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
Voltage Levels
Full Scale (FS)
InputRangeand
Programmable
Vertical Offset
50 Ω
Vertical Offset
1 MΩ
Vertical Offset
—
Range
Range
(Vpk-pk
0.04
0.1
0.2
0.4
1
)
Range (V)
(Vpk-pk
0.04
0.1
0.2
0.4
1.0
2.0
4.0
10
)
Range (V)
0.8
0.8
0.8
0.8
0.8
0.8
0.8
0.8
6.5
8.0
2
6.0
8.0
4
5.0
8.0
30
10
2.0
20
25
40
15
MaximumInput
Overload
50 Ω
1 MΩ
|Peaks| ≤35 V
—
—
7 Vrms with |Peaks| ≤10 V
Accuracy
Resolution
8 bits
DC Accuracy
(Programmable
Vertical
NI PXI-5114:
(1.5% of Input + 0.3% of FS + 200 μV)
Within 5 °C of
self-calibration
temperature
NI PCI-5114:
(1.5% of Input + 0.3% of FS + 280 μV)
Offset = 0 V)
Programmable
Vertical Offset
Accuracy
2% of offset setting
Within 5 °C of
self-calibration
temperature
DC Drift
(0.03% of Input + 0.06% of FS + 40 μV) per °C
≤–60 dB at 10 MHz
—
Crosstalk,
Typical
CH 0 to/from
CH 1, External
Trigger to CH 0
or CH 1
≤–45 dB at 100 MHz
© National Instruments Corporation
3
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
Bandwidth and Transient Response
Bandwidth
(–3 dB)
Range
Rise/FallTime,
Typical
—
(Vpk-pk
)
Bandwidth
125 MHz
All ranges
2.8 ns
except 0.04
0.04
100 MHz
3.5 ns
Bandwidth
Limit Filter
—
20 MHz Noise Filter
AC Coupling*
Cutoff (–3 dB),
Typical
*AC coupling
available on
1 MΩ only
12 Hz
Passband
Flatness
Referenced to
50 kHz
1 dB up to 50 MHz
Bandwidth
limit filter off
1
0
–1
–2
–3
–4
–5
–6
–7
–8
–9
Full Bandwidth
Noise Filter
–10
1 M
10 M
100 M
200 M
Frequency (Hz)
Figure 1. NI 5114 Frequency Response (Typical)
NI PXI/PCI-5114 Specifications
4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
Spectral Characteristics
Spurious-Free
Range (Vpk-pk
)
10 MHz,
Dynamic Range
with Harmonics
(SFDR), Typical
–1 dBFS
input signal
All ranges except 0.04
0.04
58 dBc
58 dBc
Includes the
2nd through
the 5th
Total Harmonic
Distortion
(THD), Typical
–58 dBc
–58 dBc
6.2
harmonics
Measured
from DC to
125 MHz
Effective
Number of Bits
(ENOB),
7.2
Calculated*
20 MHz
bandwidth
limit filter off
Signal to Noise
and Distortion
(SINAD),
44 dB
38 dB
Typical
RMS Noise
Range
(Vpk-pk
20 MHz
Filter On
20 MHz
Filter Off
50 Ω
)
terminator
connected
to input
All ranges
except 0.04
0.28% FS
0.28% FS
0.28% FS
0.04
0.45% FS
*ENOB = log2(sinad) – ½log2(1.5) – log2(A/V)
where
sinad = the linear representation of SINAD
A = amplitude of the supplied sine wave during the test
V = (peak) full-scale range of the waveform recorder input
Refer to 1057-1994 IEEE Standard for Digitizing Waveform Recorders for information on equation derivation.
© National Instruments Corporation
5
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Horizontal
Sample Clock
Specification
Value
Comments
Sources
Internal, Onboard Clock (internal VCXO)*
External, CLK IN (front panel SMB connector)
* Internal Sample
Clock is locked to the
Reference Clock or
derived from the
onboard VCXO
Onboard Clock (Internal VCXO)
Sample Rate
Range
Real-Time Sampling
(Single Shot)
Random Interleaved
Sampling (RIS)
* Divide by n
decimation used
for all rates less than
250 MS/s
3.815 kS/s to 250MS/s* 250 MS/s to 5 GS/s in
increments of 250 MS/s
Formoreinformation
about Sample Clock
and decimation, refer
to the NI High-Speed
Digitizers Help.
Timebase
Frequency
250 MHz
When not using
External Sample
Clock
Timebase
Accuracy
Not Phase-Locked to
Reference Clock
Phase-Locked to
Reference Clock
—
25 ppm
Equal to the Reference
Clock accuracy
Sample Clock
Delay Range
1 Sample Clock period
≤20 ps
—
—
Sample Clock
Delay Resolution
NI PXI/PCI-5114 Specifications
6
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
External Sample Clock
Sources
CLK IN (front panel SMB connector)
50 MHz to 250 MHz
—
Frequency
Range
Divide by n
decimation
available where
1 ≤ n ≤ 65,535
Formoreinformation
about Sample Clock
and decimation, refer
to the NI High-Speed
Digitizers Help.
Duty Cycle
Tolerance
45% to 55%
—
Phase-Locked Loop (PLL) Reference Clock
Specification
Value
Sources
NI PXI-5114
PXI_CLK10
NI PCI-5114
RTSI 7
CLK IN
(front panel SMB connector)
(backplane connector)
CLK IN
(front panel SMB connector)
Frequency
Range
1 MHz to 20 MHz in 1 MHz increments
Default of 10 MHz
The PLL Reference Clock frequency must be accurate to 50 ppm
45% to 55%
Duty Cycle Tolerance
Exported Reference
Clock Destinations
NI PXI-5114
NI PCI-5114
PFI <0..1> (front panel 9-pin
mini-circular DIN connector)
PFI <0..1> (front panel 9-pin
mini-circular DIN connector)
PXI_Trig <0..7> (backplane
connector)
RTSI <0..7>
© National Instruments Corporation
7
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
CLK IN (Sample Clock and Reference Clock Input,
Front Panel Connector)
Specification
Value
Input Voltage Range
Sine wave: 0.65 Vpk-pk to 2.8 Vpk-pk (0 dBm to 13 dBm)
Square wave: 0.2 Vpk-pk to 2.8 Vpk-pk
Maximum Input
Overload
7 Vrms with |Peaks| ≤10 V
Impedance
Coupling
50 Ω
AC
Trigger
Reference (Stop) Trigger
Specification
Value
Comments
Trigger Types
and Sources
Types
Sources
Refer to the
following
sections and to
NI High-Speed
Digitizers Help
for more
Edge, Window, Hysteresis,
Video, Digital, Immediate,
and Software
CH 0, CH 1, TRIG,
PXI_Trig<0..6>, PFI <0..1>,
PXI Star Trigger,
RTSI<0..6>, and Software
information.
Time
Resolution
TDC
On
Onboard Clock
External Clock
N/A
TDC = Time to
Digital
Conversion
Circuit
40 ps
4 ns
Off
External Clock Period
Rearm Time
10 μs
Rearm Time
TDC
On
Holdoff set to 0
Off
2 μs
Holdoff
From Rearm Time up to [(235 – 1) × (Sample Clock Period)]
—
—
Trigger Delay
From 0 up to [(235 – 1) – posttrigger samples] ×
(1/sample rate), in seconds
NI PXI/PCI-5114 Specifications
8
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
Analog Trigger (Edge, Window, and Hysteresis Trigger Types)
Sources
CH 0 (front panel BNC connector)
CH 1 (front panel BNC connector)
TRIG (front panel BNC connector)
8 bits (1 in 256)
—
Trigger Level
Resolution
Trigger Level
Range
CH 0, CH 1
100% FS
TRIG (External Trigger)
—
5 V
Edge Trigger
Sensitivity
5% FS up to 100 MHz
0.5 Vpk-pk up to 100 MHz
Level
5% FS up to 10 MHz
0.5 V up to 10 MHz
Accuracy,
Typical
Jitter
≤65 ps rms
—
—
Trigger Filters Low Frequency (LF) Reject High Frequency (HF) Reject
50 kHz
50 kHz
Digital Trigger (Digital Trigger Type)
Sources
NI PXI-5114
NI PCI-5114
—
PXI_Trig <0..6> (backplane RTSI <0..6>
connector)
PFI <0..1> (front panel
PFI <0..1> (front panel
SMB connector)
SMB connector)
PXI Star Trigger (backplane
connector)
Video Trigger (Video Trigger Type)
Sources CH 0 (front panel BNC connector)
CH 1 (front panel BNC connector)
TRIG (front panel BNC connector)
—
© National Instruments Corporation
9
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
Types
Specific Line
Any Line
—
Specific Field
Standards
SDTV: M-NTSC, B/G-PAL, SECAM, M-PAL
fps = fields per
second
EDTV: 480i/59.94 fps, 480i/60 fps, 480p/59.94 Fps,
480p/60 Fps, 576i/50 fps, 576p/50 Fps
Fps = Frames
per second
HDTV: 720p/50 Fps, 720p/59.94 Fps, 720p/60 Fps,
1080i/50 fps, 1080i/59.94 fps, 1080i/60 fps, 1080p/24 Fps
TRIG (External Trigger, Front Panel Connector)
Specification
Value
Connector
BNC
Impedance
1 MΩ in parallel with 22 pF
Coupling
AC, DC
12 Hz
AC-Coupling Cutoff (–3 dB)
Input Voltage Range
5 V
Maximum Input Overload
|Peaks| ≤42 V
NI PXI/PCI-5114 Specifications
10
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
PFI 0 and PFI 1 (Programmable Function Interface,
AUX Front Panel Connectors)
Specification
Connector
Value
9-pin mini-circular DIN
Bi-directional
Direction
As an Input (Trigger)
Destinations
Start Trigger (Acquisition Arm)
Reference (Stop) Trigger
Arm Reference Trigger
Advance Trigger
150 kΩ
Input Impedance
VIH
2.0 V
VIL
0.8 V
Maximum Input Overload
–0.5 V, 5.5 V
Maximum Frequency
25 MHz
As an Output (Event)
Sources
Start Trigger (Acquisition Arm)
Reference (Stop) Trigger
End of Record
Done (End of Acquisition)
Probe Compensation (1 kHz, 50% duty cycle square wave,
PFI 1 only)
Output Impedance
Logic Type
50 Ω
3.3 V CMOS
24 mA
Maximum Drive Current
Maximum Frequency
25 MHz
© National Instruments Corporation
11
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
TClk Specifications
National Instruments TClk synchronization method and the NI-TClk driver
are used to align the sample clocks on any number of SMC-based modules
in a chassis. For more information about TClk synchronization, refer to the
NI-TClk Synchronization Help, which is located within the NI High-Speed
Digitizers Help.
•
Specifications are valid for any number of modules installed in one
NI PXI-1042 chassis.
•
•
•
All parameters set to identical values for each SMC-based module.
Sample Clock set to 250 MS/s and all filters are disabled.
For other configurations, including multichassis systems, contact
NI Technical Support at ni.com/support.
Note Although you can use NI-TClk to synchronize nonidentical modules, these
specifications apply only to synchronizing identical modules.
Specification
Value
Comments
Intermodule SMC Synchronization Using NI-TClk for Identical Modules (Typical)
Skew
500 ps
Caused by clock and analog
path delay differences
No manual adjustment
performed
Skew After Manual <20 ps
Adjustment
For information about
manual adjustment, refer
to the Synchronization
Repeatability Optimization
topic in the NI-TClk
Synchronization Help.
For additional help with the
adjustment process, contact
NI Technical Support at
ni.com/support.
Sample Clock
Adjustment
Resolution
<20 ps
—
NI PXI/PCI-5114 Specifications
12
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Waveform Specifications
Specification
Value
8 megasamples
Comments
Onboard Memory
Size
8 MB per Channel
Standard
—
per channel
64 MB per Channel
Option
64 megasamples
per channel
256 MB per Channel
Option
256 megasamples
per channel
Minimum Record
Length
1 Sample
—
Number of
Pretrigger Samples
Zero up to full Record Length
Zero up to full Record Length
Single-record
mode and
multiple-record
mode
Number of
Posttrigger Samples
Single-record
mode and
multiple-record
mode
Maximum Number
of Records in
Onboard Memory
8 MB/channel
64 MB/channel
256 MB/channel
32,768
* It is possible
to exceed these
numbers if you
fetch records while
acquiring data. For
more information,
refer to the
100,000*
100,000*
NI High-Speed
Digitizers Help.
Allocated Onboard
Memory per Record
(Record Length × 1 byte/S) + 240 bytes,
rounded up to next multiple of 128 bytes
—
or
256 bytes, whichever is greater
© National Instruments Corporation
13
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration
Specification
Value
Self-Calibration
Self-calibration is done on software command. The calibration
corrects for gain, offset, compensated 1 MΩ attenuator, triggering,
and timing adjustment errors for all input ranges.
External Calibration
(Factory Calibration)
The external calibration calibrates the VCXO, gain, and the voltage
reference. Appropriate constants are stored in nonvolatile memory.
Interval for External
Calibration
2 years
Warm-Up Time
15 minutes
Power
Specification
Typical Value
NI PXI-5114
840 mA
1.1 A
NI PCI-5114
1.6 A
+3.3 VDC
+5 VDC
1.7 A
+12 VDC
–12 VDC
Total Power
250 mA
170 mA
13.32 W
45 mA
—
14.32 W
NI PXI/PCI-5114 Specifications
14
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Software
Specification
Value
NI PXI-5114: NI-SCOPE 2.9 or later
Driver Software
NI PCI-5114: NI-SCOPE 3.1 or later
NI-SCOPE is an IVI-compliant driver that allows you to configure,
control, and calibrate the NI 5114. NI-SCOPE provides application
programming interfaces for many development environments.
Application Software
NI-SCOPE provides programming interfaces, documentation, and
examples for the following application development environments:
•
•
•
•
•
LabVIEW
LabWindows™/CVI™
Measurement Studio
Microsoft Visual C/C++
Microsoft Visual Basic
Interactive Soft Front
Panel and
The Scope Soft Front Panel 2.3 or later supports interactive control of the
NI 5114. The Scope Soft Front Panel is included on the NI-SCOPE CD.
Configuration
National Instruments Measurement & Automation Explorer (MAX) also
provides interactive configuration and test tools for the NI 5114. MAX is
included on the NI-SCOPE CD.
© National Instruments Corporation
15
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Environment
NI PXI-5114
Note To ensure that the NI PXI-5114 cools effectively, follow the guidelines in
the Maintain Forced Air Cooling Note to Users included in the NI PXI-5114 kit.
The NI PXI-5114 is intended for indoor use only.
Specification
Value
0 ºC to +55 ºC in all NI PXI chassis except the following:
0 ºC to +45 ºC when installed in an NI PXI-1000/B or PXI-101x chassis
Meets IEC-60068-2-1 and IEC-60068-2-2
–40 ºC to +71 ºC
Operating Temperature
Storage Temperature
Meets IEC-60068-2-1 and IEC-60068-2-2
10% to 90%, noncondensing
Operating Relative
Humidity
Meets IEC-60068-2-56
Storage Relative
Humidity
5% to 95%, noncondensing
Meets IEC-60068-2-56
Operating Shock
30 g, half-sine, 11 ms pulse
Meets IEC-60068-2-27
Test profile developed in accordance with MIL-PRF-28800F
50 g, half-sine, 11 ms pulse
Storage Shock
Meets IEC-60068-2-27
Test profile developed in accordance with MIL-PRF-28800F
5 Hz to 500 Hz, 0.31 grms
Operating Vibration
Storage Vibration
Meets IEC-60068-2-64
5 Hz to 500 Hz, 2.46 grms
Meets IEC-60068-2-64
Test profile exceeds requirements of MIL-PRF-28800F, Class B
2,000 m maximum (at 25 °C ambient temperature)
2
Altitude
Pollution Degree
NI PXI/PCI-5114 Specifications
16
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
NI PCI-5114
Note To ensure that the NI PCI-5114 cools effectively, make sure that the chassis in which
it is used has active cooling that provides at least some airflow across the PCI card cage.
To maximize airflow and extend the life of the device, leave any adjacent PCI slots empty.
Refer to the Maintain Forced Air Cooling Note to Users included in the NI PCI-5114 kit
for important cooling information. The NI PCI-5114 is intended for indoor use only.
Specification
Value
Operating Temperature
0 ºC to +45 ºC
Meets IEC-60068-2-1 and IEC-60068-2-2
–40 ºC to +71 ºC
Storage Temperature
Meets IEC-60068-2-1 and IEC-60068-2-2
10% to 90%, noncondensing
Operating Relative
Humidity
Meets IEC-60068-2-56
Storage Relative
Humidity
5% to 95%, noncondensing
Meets IEC-60068-2-56
Storage Shock
50 g, half-sine, 11 ms pulse
Meets IEC-60068-2-27
Test profile developed in accordance with MIL-PRF-28800F
5 Hz to 500 Hz, 2.46 grms
Storage Vibration
Meets IEC-60068-2-64
Test profile exceeds requirements of MIL-PRF-28800F, Class B
2,000 m maximum (at 25 °C ambient temperature)
2
Altitude
Pollution Degree
© National Instruments Corporation
17
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Safety, Electromagnetic Compatibility,
and CE Compliance
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
•
•
IEC 61010-1, EN 61010-1
UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or visit ni.com/
certification, search by model number or product line, and click the appropriate link
in the Certification column.
Electromagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control, and
laboratory use:
•
•
•
EN 61326 EMC requirements; Minimum Immunity
EN 55011 Emissions; Group 1, Class A
CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A
Note For EMC compliance, operate this device with RG223/U or equivalent shielded
cable. Operate according to product documentation
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
•
•
2006/95/EC; Low-Voltage Directive (safety)
2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Note Refer to the Declaration of Conformity (DoC) for this product for any additional
regulatory compliance information. To obtain the DoC for this product, visit ni.com/
certification, search by model number or product line, and click the appropriate link
in the Certification column.
NI PXI/PCI-5114 Specifications
18
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Environmental Management
National Instruments is committed to designing and manufacturing
products in an environmentally responsible manner. NI recognizes that
eliminating certain hazardous substances from our products is beneficial
not only to the environment but also to NI customers.
For additional environmental information, refer to the NI and the
Environment Web page at ni.com/environment. This page contains the
environmental regulations and directives with which NI complies, as well
as any other environmental information not included in this document.
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of their life cycle, all products must be sent to a WEEE recycling
center. For more information about WEEE recycling centers and National Instruments
WEEE initiatives, visit ni.com/environment/weee.htm.
˅
RoHS
Ёᅶ᠋
݇Ѣ
DŽ
National Instruments
National Instruments
(RoHS)
Ё
ড়㾘ᗻֵᙃˈ䇋ⱏᔩ ni.com/environment/rohs_chinaDŽ
RoHS
ni.com/environment/rohs_china
(For information about China RoHS compliance, go to
.)
Physical
Front Panel Connectors
Label
CH 0
Function
Connector Type
BNC female
Comments
—
Analog Input
Analog Input
External Trigger
CH 1
BNC female
BNC female
SMB jack
TRIG
CLK IN
Sample Clock Input and
Reference Clock Input
AUX I/O
PFI 0, PFI 1
9-pin mini-circular DIN
© National Instruments Corporation
19
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
NI PXI-5114 Front Panel Indicators
Label
Function
For more
information,
refer to the
NI High-Speed
Digitizers Help.
ACCESS
The ACCESS LED indicates the status of the PCI bus and
the interface from the NI PXI-5114 to the controller.
ACTIVE
The ACTIVE LED indicates the status of the onboard
acquisition hardware of the NI PXI-5114.
Dimensions and Weight
NI PXI-5114
Dimensions
3U, One slot, PXI/cPCI Module
21.6 × 2.0 × 13.0 cm
(8.5 × 0.8 × 5.1 in.)
2.0 cm
(0.8 in.)
13.0 cm
(5.1 in.)
21.6 cm
(8.5 in.)
Weight
455 g (16 oz)
NI PXI/PCI-5114 Specifications
20
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
NI PCI-5114
Dimensions
35.5 × 2.0 × 11.3 cm
(14.0 × 0.8 × 4.4 in.)
2.0 cm
(0.8 in.)
11.3 cm
(4.4 in.)
35.5 cm
(14.0 in.)
Weight
421 g (14.8 oz)
© National Instruments Corporation
21
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Where to Go for Support
The National Instruments Web site is your complete resource for technical
support. At ni.com/support you have access to everything from
troubleshooting and application development self-help resources to email
and phone assistance from NI Application Engineers.
A Declaration of Conformity (DoC) is our claim of compliance with the
Council of the European Communities using the manufacturer’s
declaration of conformity. This system affords the user protection for
electronic compatibility (EMC) and product safety. You can obtain the DoC
for your product by visiting ni.com/certification. If your product
supports calibration, you can obtain the calibration certificate for your
product at ni.com/calibration.
National Instruments corporate headquarters is located at
11500 North Mopac Expressway, Austin, Texas, 78759-3504.
National Instruments also has offices located around the world to help
address your support needs. For telephone support in the United States,
create your service request at ni.com/support and follow the calling
instructions or dial 512 795 8248. For telephone support outside the United
States, contact your local branch office:
Australia 1800 300 800, Austria 43 662 457990-0,
Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599,
Canada 800 433 3488, China 86 21 5050 9800,
Czech Republic 420 224 235 774, Denmark 45 45 76 26 00,
Finland 358 (0) 9 725 72511, France 01 57 66 24 24,
Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737,
Italy 39 02 413091, Japan 81 3 5472 2970, Korea 82 02 3451 3400,
Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710,
Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466,
New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60,
Poland 48 22 3390150,Portugal 351 210 311 210,Russia 7 495 783 6851,
Singapore 1800 226 5886, Slovenia 386 3 425 42 00,
South Africa 27 0 11 805 8197, Spain 34 91 640 0085,
Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151,
Taiwan 886 02 2377 2222, Thailand 662 278 6777,
Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545
NI PXI/PCI-5114 Specifications
22
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation.
Refer to the Terms of Use section on ni.com/legal for more information about National
Instruments trademarks. Other product and company names mentioned herein are trademarks or trade
names of their respective companies. For patents covering National Instruments products, refer to the
appropriate location: Help»Patents in your software, the patents.txt file on your CD, or
ni.com/patents.
© 2005–2007 National Instruments Corporation. All rights reserved.