Agilent Technologies Respiratory Product 54600 Series User Manual |
Agilent Technologies
54600-Series Oscilloscopes
Data Sheet
Easily see what's happening in your
mixed analog and digital designs
Multiple configurations to
meet your needs
•
•
60 to 500 MHz, up to 2 GSa/s
If you work with both analog
and digital components, Agilent
Technologies 54600-Series
• A revolutionary ultra-respon-
sive, high-definition display
that lets you see more signal
detail than ever before
Unique 2 + 16-channel MSO
and 2- or 4-channel models
oscilloscopes can help you easily
see more of what's going on in
your designs. The unique 2 +
16-channel mixed signal oscillo-
scope (MSO) models and the tra-
ditional 2- and 4-channel models
are optimized with just the capa-
bilities you need for verifying and
debugging designs that include
A/Ds, D/As, DSPs, and embedded
8- or 16-bit microcontrollers.
These scopes give you the tools
you need to solve your mixed
analog and digital engineering
challenges more easily.
•
•
•
•
4 MB to 8 MB MegaZoom deep
memory standard
• Flexible triggering that lets you
easily isolate and analyze the
complex signals and fault
Patented high-definition
display system
Flexible triggering including CAN,
conditions common in mixed
analog and digital designs.
2
I C, LIN, SPI, and USB
2
I C, CAN, LIN, USB, and SPI
Deep memory transfer over
the interface bus
triggering come standard.
N2758A CAN trigger module
option is also available.
This combination of capabilities
is tailored to give you the meas-
urement power you need to get
your mixed analog and digital job
done faster.
Ideal for mixed analog and
digital analysis
At Agilent Technologies, we focus
on developing products that help
you do your job better. That’s
why 54600-Series scopes are
optimized for your needs. Choose
the one that’s right for your
Ideal for analyzing designs with
both analog and digital com-
ponents, the 54600-Series scopes
combine three critical features:
application and your budget.
• Up to 8 MB of MegaZoom deep
memory come standard so you
can capture long, non-repeat-
ing signals, maintain high sam-
ple rate, and quickly zoom in
on areas of interest
Download from Www.Somanuals.com. All Manuals Search And Download.
Selection Guide (continued)
Figure 2. 2 + 16-channel mixed signal oscilloscope (MSO) allows you to view analog and digital content on one instrument.
4-channel scope
(model 54624A)
2-channel scopes
(models 54621A, 54622A,
54641A, and 54642A)
Mixed-signal scopes
(models 54621D, 54622D,
54641D, and 54642D)
If your designs include heavy
analog content, the 100-MHz
54624A will give you the channel
count and measurement power
you need, including MegaZoom
deep memory, high-definition
display, and flexible triggering.
Whether you’re testing designs
with four inputs, such as anti-lock
brakes, or monitoring multiple
outputs of a power supply, the
4-channel model helps you get
your debug and verification done
with ease.
The mixed-signal oscilloscopes
(MSO), with 2 analog channels
and 16 digital channels, uniquely
combine the detailed signal analy-
sis of a scope with the multi-
channel timing measurements of
a logic analyzer. They let you see
the complex interactions among
your signals on up to 18 channels
at the same time. No more guess-
work and no more poking around
to see a few channels at a time.
These scopes can easily conquer
mixed analog and digital debug-
ging problems that a traditional
scope can't begin to address,
because they let you simultane-
ously test and monitor the high-
speed digital control signals and
the slower analog signals in
The 2-channel models bring all
the benefits of MegaZoom deep
memory, high-definition display,
and flexible triggering to those
value-minded designers with
lower channel requirements. They
give you an affordable way to see
long time periods while maintain-
ing high sample rate so you can
see details in your designs.
your design.
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Affordable and Responsive
MegaZoom deep memory
With 4 MB to 8 MB of MegaZoom
deep memory, 54600-Series
scopes give you deep-memory
capture without the sluggish
response and complex operation
you’ve had to tolerate with other
deep-memory scopes. And unlike
the alternatives, MegaZoom deep
memory is not a special mode.
That means it’s always available
to help you find details buried
in complex signals, discover
anomalies, correlate high-speed
digital control signals with slower
analog signals and capture
infrequent events.
The dilemma of having to trigger
twice to get a long time capture
or to see detailed resolution is
solved with MegaZoom deep
memory as you can have both
with a single measurement. Deep
memory means that the sample
rate stays high even when captur-
ing long time periods.
The 54600-Series scopes are the
only deep-memory scopes that
respond instantly to your control
inputs with a fast, responsive
display. Compare them to other
scopes in this price range – only
the 54600-Series provides deep
memory at an affordable price.
Figure 3. The bright dot on the high-definition display is a distortion in 1 of the 1,500
pulses captured in this single-shot measurement. Simply dial in for a closer look using
MegaZoom pan and zoom, and you’ll see the details that would have escaped other
scopes, such as the distortion in this square wave. This deep memory and display system
is not a special mode; it is available on every measurement pass at full speed.
4
Download from Www.Somanuals.com. All Manuals Search And Download.
Affordable and Responsive (continued)
For more information on
IntuiLink, please see the
Revolutionary high-definition dis-
play helps you see details
Connectivity made easier
IntuiLink datasheet with
Agilent publication number
5980-3115EUS or visit the URL
• Parallel and RS-232 interfaces
make connection to printers
and PCs a snap. And the
Combine MegaZoom deep memory
with a patented high-definition
display system, and you will see
subtle details that most scopes
won’t show you. The MegaZoom
deep memory is mapped into 32
levels of intensity on a display
that has twice the horizontal
resolution and update rates of up
to 25 million vectors per second –
so you see a more realistic
54600-Series scopes come with
standard parallel and RS-232
interfaces on the rear panel.
For faster data transfers, an
optional GPIB interface
Built-in floppy drive
A built-in 1.44 MB floppy drive
makes it easy for you to store
waveform data, screen images,
and scope setups. You can store
your waveform images as TIF or
BMP files and your waveform
data as ASCII files for easy
import into other PC applica-
tions. If you share your lab equip-
ment with others, you can save
your measurement setups and
traces to diskette, making it sim-
ple to reproduce your every-day
setups as well as your advanced
configurations.
module is also available.
• IntuiLink, a free software applica-
tion, simplifies PC connectivity
when you need to transfer
images and waveform data to
your PC. IntuiLink lets you
focus more time on design
issues rather than on program-
ming. With the click of a but-
ton on IntuiLink’s toolbar, you
can download data or insert a
snapshot of an oscilloscope
screen directly into your
representation of your signals.
There is less chance to miss a
narrow occasional transient, less
chance to miss a glitch or distort-
ed edge that impacts circuit
operation, and less chance to
miss all those subtle details that
can take weeks to find with a
traditional digital scope.
Microsoft® Excel spreadsheets
or Word documents, or save
the image as a bitmap file.
Once the data is in an Excel
spreadsheet, you can leverage
Excel’s extensive analysis
functions to create custom
graphs and views.
Built-in help at your fingertips
Standard features
An innovative built-in help
system in 11 different languages
(English, French, German,
Agilent 54600-Series scopes
include the standard features you
need to get your job done easier
and faster.
Spanish, Italian, Japanese,
Korean, Traditional Chinese,
Simplified Chinese, Russian, and
Portuguese) gives you quick
access to the help you need. If
you have a question about a par-
ticular feature, simply press and
hold the corresponding front-
panel key for a few seconds, and
a help screen pops up to explain
its function. You’ll no longer have
to look for the manual when you
need assistance setting up scope
functions or making complex
measurements.
Figure 4. Press and hold any key for built-in Help, such as this
description of the frequency measurement.
5
Download from Www.Somanuals.com. All Manuals Search And Download.
Affordable and Responsive (continued)
Deep memory transfer over
the bus
Other standard features
• Waveform math with FFTs:
Analysis functions include -, *,
Integrate, and Differentiate,
as well as Fast Fourier
Transforms.
Transfer all of your data over the
bus, even when using deep
memory. Typical transfer time
is <70 seconds for 2 MB of data
over GPIB.
• 1-ns peak detect on the
54640-Series and 5-ns peak
detect on the 54620-Series
means you won’t have to worry
about missing narrow glitches.
The 54640-Series requires soft-
ware revision 1.0 or later. The
54620-Series requires software
revision 1.5 or later. The system
software is stored in FLASH ROM
that can be upgraded easily from
a built-in floppy disk drive in the
scope. Visit the 54620-Series
customer web site at:
for a free copy of the system
software and the IntuiLink Data
Capture application to pull the
data out of the scope to the PC.
• Autoscale lets you quickly dis-
play any active signals, auto-
matically setting the vertical,
horizontal and trigger controls
for the best display.
• Printer connectivity is easy
for HP Deskjets, Laserjets,
Epson, or integrated thermal
printer option with the stan-
dard parallel port on the rear
panel.
• High resolution mode offers
up to 12 bits of resolution in
real-time (single-shot) modes,
reducing noise. This is accom-
plished by averaging sequential
data points and mapping the
average to the display when at
slow sweep speeds.
• 1-year warranty: All
54600-Series scopes include a
full 1-year warranty with
optional 3-year and 5-year
warranty coverage.
• Money-back guarantee: Get a
full refund if you’re not satis-
fied with your purchase for
any reason. See page 24 for
more information.
6
Download from Www.Somanuals.com. All Manuals Search And Download.
N2758A CAN Trigger Module
•
Trigger on CAN 2.0A/B message
frames with user specified IDs
and/or data
•
•
•
•
•
•
Trigger on Error frames
Differential CAN
Acknowledge On/Off
Built In QuickHelp
Remote Control via GPIB/RS232
Compatible with 54621D, 54622D,
54641D, 54642D
The N2758A CAN trigger module
extends your Mixed Signal
Oscilloscope’s (MSO) capabilities
beyond the standard CAN SOF
trigger. The module enables the
MSO to isolate and trigger on a
particular frame’s content. This
triggering allows you to:
The N2758A kit includes the CAN trigger module, digital cable, ten grabbers, and
two probe ground leads.
• sort through frame traffic on
the bus
• isolate frames of interest
• specify particular frame
characteristics
• trigger and measure latency
between CAN frames meeting
the user specification
The CAN trigger module is ideal
for analyzing all traffic on the
broadcast network. Quickly find
frames that meet the specifica-
tions you select with an isolating
capability similar to pattern
triggering across multiple
channels in the parallel domain.
Synchronize to the particular
frame then examine the system
behavior of interest.
For more comprehensive
information, please see the
Agilent 54600-Series
Oscilloscopes Probes and
Accessories data sheet (Agilent
publication number 5968-8153).
You can get a copy by calling
1-800-452-4844 or by visiting
our website at:
The N2758A CAN trigger module reduces the time required to
debug your mixed signal CAN based designs.
7
Download from Www.Somanuals.com. All Manuals Search And Download.
Triggering
Figure 6. Make your automotive and industrial testing easier.
CAN triggering synchronizes to the start of a CAN frame on any
CAN high or CAN low signal.
Figure 5. Easily synchronize to a
USB packet.
USB is quickly becoming the
Flexible Triggering
CAN, LIN, and USB Triggering
connection of choice for PC
peripherals. Make your testing
easy with Start of Packet (SOP),
End of Packet (EOP), Reset
Complete (RC), Enter Suspend,
or Exit Suspend modes.
With mixed analog and digital
designs, sometimes it is hard to
trace an anomaly back to its root
cause unless you can trigger on it
and correlate it with another
signal. With the 54600-Series
scopes, triggering is no longer a
hassle. They come with flexible
triggering capabilities across all
channels so you can easily isolate
and analyze complex signals and
interactions common in your
mixed analog and digital designs.
Controller Area Network (CAN)
and Universal Serial Bus (USB)
triggering comes standard with
the 54600-Series scopes. These
triggering modes make it easy to
synchronize CAN, LIN, and USB
frames. Don’t waste time finding
the start of the complex frames –
let Agilent do it for you.
The N2758A CAN trigger module
option enables the MSO to isolate
and trigger on a particular
frame’s content. For more
information , see page 7.
CAN triggering synchronizes to
the start of a CAN frame on any
CAN high or CAN low signal.
Trigger on LIN (local interconnect
network) sync break at the begin-
ning of a message frame.
8
Download from Www.Somanuals.com. All Manuals Search And Download.
Triggering (continued)
2
2
The full range of I C triggering
I C and SPI Triggering
includes start or stop condition,
missing acknowledge, restart,
EEPROM data read, address
and/or data frame, or 10-bit
write. SPI triggering allows for
trigger on user-definable framing
and user-definable number of
bits per frame, as well as data
patterns.
Don’t spend your time sorting
through communication frames to
find the one of interest. Let
Agilent set up an Inter-Integrated
2
Circuit (I C) or Serial Peripheral
Interface (SPI) trigger to sift
through the frames for you. And
like the CAN, LIN, and USB trig-
2
gers, I C and SPI come standard
with the 54600-Series scopes.
Figure 8. Trigger on address and/or data
2
patterns within an I C frame.
Figure 7. Trigger on data patterns within a SPI frame.
quate when it comes to the chan-
nel count, memory depth, display
and triggering. The interactions
that occur in typical mixed analog
and digital design — taking a real-
world signal such as audio, video,
temperature or pressure, per-
forming some transformation on
the signal, and finally returning
that signal to the real world —
require more viewing capability
than current scopes can provide.
Microcontroller and
DSP Applications
Microcontrollers (MCUs) and
Digital Signal Processors (DSPs)
are here to stay, and are becoming
more pervasive in industrial,
communications, automobile, and
consumer products. Design engi-
neers must grapple with scopes
that were not developed to tackle
the mixed analog and digital
signals found in MCU- and
Agilent 54621D, 54622D, 54641D,
and 54642D mixed-signal
DSP-based designs. Many engi-
neers, who rely primarily on
typical 2-channel or 4-channel
scopes in mixed analog and digi-
tal signal verification and debug,
find these scopes to be inade-
oscillscopes (MSOs) provide the
channel count, memory depth,
high-definition display, and
triggering that MCU and DSP
designs demand.
Figure 9. The 350 MHz and 500 MHz MSOs
are a great fit for DSP applications.
9
Download from Www.Somanuals.com. All Manuals Search And Download.
MegaZoom deep memory helps
you determine how your signals
are impacting each other. With
shallow memory scopes, you have
to choose whether you look at a
slow analog signal or fast digital
content. With up to 8 MB deep
memory, you don’t have to choose
– capture all of your data at once.
Revolutionary high-definition
display reveals subtle details that
most scopes won’t show you.
Standard RS-232 and parallel
ports provide PC and printer
connectivity.
Free IntuiLink PC software
makes transfer of waveform data
to PC fast and easy.
Intensity knob allows you to
adjust the contrast on screen.
Built-in floppy drive makes it
easy to save your work and
update your system software.
10
Download from Www.Somanuals.com. All Manuals Search And Download.
Quickly pan and zoom for
analysis with MegaZoom’s instant
response and optimum resolution.
Autoscale lets you quickly
display any active signals,
automatically setting the vertical,
horizontal and trigger controls
for the best display, while opti-
mizing memory.
QuickMeas gives you any three
automated measurements with
the push of a button. QuickPrint
automatically prints your screen
or saves to disk with automated
file names.
Standard Serial Triggering
2
includes CAN, I C, LIN, SPI,
and USB
Built-in help is available in 11
languages. Simply press and hold
the front-panel key of interest for
a few seconds, and a help screen
pops up to explain its function.
Maximum sample rate and
resolution on every measure-
ment. The scope automatically
adjusts memory depth as you use
it, so you get maximum sample
rate and resolution on every
measurement. You don’t even
have to think about it.
2 analog and 16 digital channel
MSOs allow you to see up to 18
time-aligned signals on your
scope screen. Also available in
2- and 4-channel models.
11
Download from Www.Somanuals.com. All Manuals Search And Download.
Probes
To get the most out of your scope,
you need the right probes and
accessories for your particular
application. That's why Agilent
Technologies offers a complete
family of innovative probes and
54600-Series scopes. For more
comprehensive information,
please see the Agilent
54600-Series Oscilloscopes
Probes and Accessories
data sheet (Agilent publication
number 5968-8153). You can get a
copy by calling 1-800-452-4844 or
by visiting our website at:
Agilent Passive Probes Selection Guide
10070C
10073C (shipped
10074C (shipped
10076A
N2771A
with 54640-Series)
with 54620-Series)
high-voltage probe
high-voltage probe
Probe Bandwidth
20 MHz
< 17.5 ns
1:1
500 MHz
< 700 ps
10:1
150 MHz
< 2.33 ns
10:1
250 MHz
< 1.4 ns
100:1
50 MHz
< 7 ns
Probe Risetime (Calculated)
Attenuation Ratio
1000:1
100 MΩ
Input Resistance
1 MΩ
2.2 MΩ
10 MΩ
66.7 MΩ
(when terminated into 1 MΩ)
Input Capacitance
Approx. 70 pF
Approx. 12 pF
Approx. 15 pF
Approx. 3 pF
4000 Vpk
Approx. 1 pF
Maximum Input
(dc + peak ac)
400 Vpk CAT I
(mains isolated)
400 Vpk CAT II
(post receptacle
mains)
500 Vpk CAT I
(mains isolated)
400 Vpk CAT II
(post receptacle
mains)
500 Vpk CAT I
(mains isolated)
400 Vpk CAT II
(post receptacle
mains)
15 kV dc, 10 kVrms
30 kV dc + peak ac
Compensation Range
Probe Sense
None
No
6 - 15 pF
Yes
9 - 17 pF
Yes
6 - 20 pF
Yes
7 - 25 pF
No
Current Probes
Description
1146A
100 kHz current probe, ac/dc
50 MHz current probe, ac/dc
power supply for N2774A
N2774A
N2775A
Differential Probes
Description
N2772A
N2773A
20 MHz differential probe
Differential probe power supply
12
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Characteristics
Acquisition: Analog Channels
Max Sample Rate
54621A/D, 54622A/D, 54624A: 200 MSa/s
54641A/D, 54642A/D: 2 GSa/sec interleaved, 1 GSa/sec each channel
Max Memory Depth
54621A/D, 54622A/D, 54624A: 4 MB interleaved, 2 MB each channel
54641A/D, 54642A/D: 8 MB interleaved, 4 MB each channel
Vertical Resolution
Peak Detection
8 bits
54621A/D, 54622A/D, 54624A: 5 ns
54641A/D, 54642A/D: 1 ns @ max sample rate
Averaging
Selectable from 2, 4, 8, 16, 32, 64 … to 16383
High Resolution Mode
54621A/D, 54622A/D, 54624A: 12 bits of resolution when ≥500 µs/div, (average mode with ave = 1)
54641A/D, 54642A/D: 12 bits of resolution when ≥100 µs/div, (average mode with ave = 1)
Filter
Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of scope, whichever is less) with vectors on.
Acquisition: Digital Channels (54621D, 54622D, 54641D, and 54642D only)
Max Sample Rate
Max Memory Depth
Vertical Resolution
54621D, 54622D: 400 MSa/s interleaved, 200 MSa/s each channel
54641D, 54642D: 1 GSa/s
54621D, 54622D: 8 MB interleaved, 4 MB ea. channel
54641D, 54642D: 4 MB
1 bit
5 ns
Glitch Detection
(min pulse width)
Vertical System: Analog Channels
Analog Channels
54621A/D, 54622A/D, 54641A/D, 54642A/D: Ch 1 and 2 simultaneous acquisition
54624A: Ch 1, 2, 3 and 4 simultaneous acquisition
Bandwidth (-3dB)*
ac Coupled
54621A/D: dc to 60 MHz
54622A/D, 54624A: dc to 100 MHz
54641A/D: dc to 350 MHz
54642A/D: dc to 500 MHz
54621A/D: 3.5 Hz to 60 MHz
54622A/D: 54624A: 3.5 Hz to 100 MHz
54641A/D: 3.5 Hz to 350 MHz
54642A/D: 3.5 Hz to 500 MHz
Calculated Risetime
(=0.35/bandwidth)
54621A/D: ~5.8 ns
54622A/D, 54624A: ~3.5 ns
54641A/D: ~1.0 ns
54642A/D: ~700 ps
Single Shot Bandwidth
54621A/D, 54622A/D, 54624A: 50 MHz
54641A/D: 350 MHz maximum
54642A/D: 500 MHz maximum
*
Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and 10 °C from firmware calibration temperature.
13
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Characteristics (continued)
Vertical System: Analog Channels (continued)
1,2
Range
54621A/D, 54622A/D, 54624A: 1 mV/div to 5 V/div
54641A/D, 54642A/D: 2 mV/div to 5 V/div
Maximum Input
CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk
With 10073C/10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50 Ω input
Offset Range
54621A/D, 54622A/D, 54624A:
5 V on ranges <10 mV/div; 25 V on ranges 10 mV/div to 199 mV/div; 100 V on ranges ≥200 mV/div
54641A/D, 54642A/D:
5 V on ranges <10 mV/div; 20 V on ranges 10 mV/div to 200 mV/div; 75 V on ranges >200 mV/div
Dynamic Range
Input Resistance
Lesser of 8 div or 32 V from center screen
54621A/D, 54622A/D, 54624A: 1 MΩ 1%
54641A/D, 54642A/D: 1 MΩ 1%, 50 Ω selectable
Input Capacitance
Coupling
54621A/D, 54622A/D, 54624A: ~ 14 pF
54641A/D, 54642A/D: ~ 13 pF
54621A/D, 54622A/D, 54624A: ac, dc, ground
54641A/D, 54642A/D: ac, dc
BW Limit
54621A/D, 54622A/D, 54624A: ~ 20 MHz selectable
54641A/D, 54642A/D: ~ 25 MHz selectable
Channel-to-Channel Isolation
(with channels at same V/div)
54621A/D, 54622A/D, 54624A: dc to 20 MHz > 40 dB; 20 MHz to max bandwidth > 30 dB
54641A/D, 54642A/D:DC to max bandwidth >40 dB
Probes
54621A/D, 54622A/D, 54624A: 10:1 10074C shipped standard for each analog channel
54641A/D, 54642A/D: 10:1 10073C shipped standard for each analog channel
Probe ID (Agilent/HP and
Tek compatible)
Auto probe sense
2 kV
ESD Tolerance
Noise Peak-to-Peak
54621A/D, 54622A/D, 54624A: 2% full scale or 1 mV, whichever is greater
54641A/D, 54642A/D: 3% full scale or 3 mV, whichever is greater
Common Mode Rejection Ratio 20 dB @ 50 MHz
1,2
DC Vertical Gain Accuracy*
2.0% full scale
DC Vertical Offset Accuracy
54621A/D, 54622A/D, 54624A:
<200 mV/div: 0.1 div 1.0 mV 0.5% offset value; ≥200 mV/div: 0.1 div 1.0 mV 1.5% offset value
54641A/D, 54642A/D:
≤200 mV/div: 0.1 div 2.0 mV 0.5% offset value; >200 mV/div: 0.1 div 2.0 mV 1.5% offset value
1,2
Single Cursor Accuracy
{DC Vertical Gain Accuracy + DC Vertical Offset Accuracy + 0.2% full scale (~1/2 LSB) }
54621A/D, 54622A/D, 54624A example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = {2.0%(80mV) + 0.1 (10 mV) + 1.0 mV + 0.5% (5 mV) + 0.2%(80 mV)} = 3.78 mV
1,2
Dual Cursor Accuracy*
{DC Vertical Gain Accuracy + 0.4% full scale (~1 LSB)}
Example: for 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = {2.0%(80 mV) + 0.4%(80 mV)} = 1.92 mV
*
1
Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and 10 °C from firmware calibration temperature.
For 54621A, 54621D, 54622A, 54622D, and 54624A, 1 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for
1 mV/div sensitivity setting.
2
For 54641A, 54641D, 54642A, and 54642D, 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 2 mV/div sensitivity setting.
14
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Characteristics (continued)
Vertical System: Digital Channels (54621D, 54622D, 54641D, and 54642D only)
Number of Channels
Threshold Groupings
16 Digital – labeled D15 - D0
Pod 1: D7 - D0
Pod 2: D15 - D8
Threshold Selections
TTL, CMOS, ECL, user-definable (selectable by pod)
8.0 V in 10 mV increments
40 V peak CAT I
User-Defined Threshold Range
Maximum Input Voltage
Threshold Accuracy*
(100 mV + 3% of threshold setting)
10 V about threshold
Input Dynamic Range
Minimum Input Voltage Swing
Input Capacitance
500 mV peak-to-peak
~ 8 pF
Input Resistance
100 kΩ 2% at probe tip
2 ns typical, 3 ns maximum
Channel-to-Channel Skew
Horizontal
Range
54621A/D, 54622A/D, 54624A: 5 ns/div to 50 s/div
54641A/D, 54642A/D: 1 ns/div to 50 s/div
Resolution
54621A/D, 54622A/D, 54624A: 25 ps
54641A/D, 54642A/D: 2.5 ps
Vernier
1-2-5 increments when off, ~25 minor increments between major settings when on
Left, Center, Right
Reference Positions
Delay Range
54621A/D, 54622A/D, 54624A:
Pre-trigger (negative delay): Greater of 1 screen width or 10 ms
Post-trigger (positive delay): 500 seconds
54641A/D, 54642A/D:
Pre-trigger (negative delay): Greater of 1 screen width or 1 ms
Post-trigger (positive delay): 500 seconds
Analog Delta-t Accuracy
54621A/D, 54622A/D, 54624A:
Same Channel*: 0.01% reading 0.1% screen width 40 ps
Channel-to-Channel: 0.01% reading 0.1% screen width 80 ps
54641A/D, 54642A/D:
Same Channel*: 0.005% reading 0.1% screen width 20 ps
Channel-to-Channel: 0.005% reading 0.1% screen width 40 ps
Same Channel Example (54641A/D, 54642A/D):
for signal with pulse width of 10 µs, scope set to 5 µs/div (50 µs screen width),
delta-t accuracy = {.005%(10 µs) + 0.1%(50 µs) + 20 ps} = 50.52 ns
*
Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and 10 °C from firmware calibration temperature.
15
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Characteristics (continued)
Horizontal (continued)
Digital Delta-t Accuracy
(non-Vernier settings)
54621A/D, 54622A/D, 54624A:
Same Channel: 0.01% reading 0.1% screen width (1 digital sample period, 2.5 or 5 ns
based on sample rate of 200/400 MSa/s)
Channel-to-Channel: 0.01% reading 0.1% screen width (1 digital sample period, 2.5 or 5 ns)
chan-to-chan skew (2 ns typical, 3 ns maximum)
54641A/D, 54642A/D:
Same Channel: 0.005% reading 0.1% screen width (1 digital sample period, 1ns)
Channel-to-Channel: 0.005% reading 0.1% screen width (1 digital sample period) chan-to-chan skew
Same Channel Example (54641A/D, 54642A/D):
for signal with pulse width of 10 µs, scope set to 5 µs/div (50 µs screen width),
delta-t accuracy = {.005%(10 µs) + 0.1%(50 µs) + 1 ns} = 51.5 ns
Delay Jitter
RMS Jitter
Modes
<1 ppm
0.025% screen width + 30 ps
Main, Delayed, Roll, XY
XY
Bandwidth: Max bandwidth
Phase error @ 1 MHz: 1.8 degrees
Z Blanking: 1.4 V blanks trace (use external trigger) - 54621A/D, 54622A/D, 54624A only
Trigger System
Sources:
54621A/622A, 54641A/642A: Ch 1, 2, line, ext
54621D/622D, 54641D/642D: Ch 1, 2, line, ext, D15 - D0
54624A: Ch 1, 2, 3, 4, line, ext
Modes
Auto, Triggered (normal), SingleAuto level (54621A/D, 54622A/D, 54624A only)
~60 ns to 10 seconds
Holdoff Time
Selections
Edge
2
Edge, Pulse Width, Pattern, TV, Duration, Sequence, CAN, LIN, USB, I C, SPI
Trigger on a rising or falling edge of any source
Pattern
Trigger on a pattern of high, low, and don’t care levels and/or a rising or falling edge established across any of the sources.
The analog channel’s high or low level is defined by that channel’s trigger level.
Pulse Width
Trigger when a positive- or negative-going pulse is less than, greater than, or within a specified range on any of
the source channels.
Minimum pulse width setting: 5ns (2 ns on 54641A/D, 54642A/D analog channels)
Maximum pulse width setting: 10 s
TV
Trigger on any analog channel for NTSC, PAL, PAL-M, or SECAM broadcast standards on either positive or negative
composite video signals. Modes supported include Field 1, Field 2, or both, all lines, or any line within a field. Also supports
triggering on non-interlaced fields. TV trigger sensitivity: 0.5 division of synch signal.
Sequence
CAN
Arm on event A, trigger on event B, with option to reset on event C or time delay.
Trigger on CAN (Controller Area Network) version 2.0A and 2.0B signals. It can trigger on the Start of Frame bit of a data
frame, a remote transfer request frame, or an overload frame.
LIN
Trigger on LIN (Local Interconnect Networking) sync break at beginning of message frame.
USB
Trigger on USB (Universal Serial Bus) Start of Packet, End of Packet, Reset Complete, Enter Suspend, or Exit Suspend on
the differential USB data lines. USB low speed and full speed are supported.
*
Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and 10 °C from firmware calibration temperature.
16
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Characteristics (continued)
Trigger System (continued)
2
I C
2
Trigger on I C (Inter-IC bus) serial protocol at a start/stop condition or user defined frame with address and/or data
values. Also trigger on Missing Acknowledge, Restart, EEPROM read, and 10 bit write.
SPI
Trigger on SPI (Serial Protocol Interface) data pattern during a specific framing period. Support positive and negative Chip
Select framing as well as clock Idle framing and user specified number of bits per frame.
Duration
Trigger on a multi-channel pattern whose time duration is less than a value, greater than a value, greater than a time value
with a timeout value, or inside or outside of a set of time values.
Minimum duration setting: 5 ns
Maximum duration setting: 10 s
Autoscale
Finds and displays all active analog and digital (for 54621D/54622D/54641D/54642D) channels, sets edge trigger mode on
highest numbered channel, sets vertical sensitivity on analog channels and thresholds on digital channels, time base to
display ~1.8 periods. Requires minimum voltage >10 mVpp, 0.5% duty cycle and minimum frequency >50Hz.
Analog Channel Triggering
Range (internal)
6 div from center screen
Sensitivity*
54621A/D, 54622A/D, 54624A: Greater of 0.35 div or 2.5 mV
54641A/D, 54642A/D:<10mV/div: greater of 1 div or 5mV; ≥10mV/div: 0.6 div
Coupling
AC (~3.5 Hz on 54621A/D, 54622A/D, 54624A.
~10 Hz on 54641A/D, 54642A/D), DC, noise reject, HF reject and LF reject (~ 50 kHz)
Digital (D15 - D0) Channel Triggering (54621D, 54622D, 54641D, and 54642D)
Threshold Range (used defined)
Threshold Accuracy*
8.0 V in 10 mV increments
(100 mV + 3% of threshold setting)
TTL = 1.4 V, CMOS = 2.5 V, ECL = -1.3 V
Predefined Thresholds
External (EXT) Triggering
Input Resistance
54621A/D, 54622A/D, 54624A: 1 MΩ, 3%
54641A/D, 54642A/D: 1 MΩ 3% or 50 Ω
Input Capacitance
Maximum Input
54621A/D, 54622A/D, 54624A: ~ 14 pF
54641A/D, 54642A/D: ~13pF
CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk
With 10073C/10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50-ohm input
Range
54621A/D, 54622A/D, 54624A: 10 V
54641A/D, 54642A/D: DC coupling: trigger level 8V; AC coup./LFR: AC input minus trig level not to exceed 8V
Sensitivity
Coupling
54621A/D, 54622A/D, 54624A: dc to 25 MHz, < 75 mV; 25 MHz to max bandwidth, < 150 mV
54641A/D, 54642A/D: dc to 100 MHz, < 100 mV; 100 MHz to max bandwidth, < 200 mV
AC (~ 3.5 Hz), DC, noise reject, HF reject and LF reject (~ 50 kHz)
Auto probe sense for 54621A/622A/641A/642A
Probe ID (Agilent/HP and
Tek compatible)
*
Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and 10 °C from firmware calibration temperature.
17
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Characteristics (continued)
Display System
Display
7-inch raster monochrome CRT
Throughput of Analog Channels 25 million vectors/sec per channel with 32 levels of intensity
Resolution
255 vertical by 1000 horizontal points (waveform area) 32 levels of gray scale
Waveform intensity on front panel. Vectors on/off; infinite persistence on/off 8 x 10 grid with continuous intensity control
Key-specific help in 11 languages displayed by pressing and holding key or softkey of interest
Time and date (user setable)
Controls
Built-in Help System
Real Time Clock
Measurement Features
Automatic Measurements
Voltage (analog channels only)
Time
Measurements are continuously updated. Cursors track current measurement
Peak-to-Peak, Maximum, Minimum, Average, Amplitude, Top, Base, Overshoot, Preshoot, RMS (DC)
Frequency, Period, + Width, - Width and Duty Cycle on any channels
Rise time, Fall time, X at Max (Time at max volts), X at Min (Time at min volts), Delay, and Phase on analog channels only.
Counter
Built-in 5 digit frequency counter on any channel. Counts up to 125 MHz
Threshold Definition
Cursors
Variable by percent and absolute value; 10%, 50%, 90% default for time measurements
Manually or automatically placed readout of Horizontal (X, ∆X, 1/∆X) and Vertical (Y, ∆Y)
Additionally digital or analog channels can be displayed as binary or hex values
Waveform Math
One function of 1-2, 1*2, FFT, differentiate, integrate.
Source of FFT, differentiate, integrate: analog channels 1 or 2 (or 3 or 4 for 54624A), 1-2, 1+2, 1*2
FFT
Points
Fixed at 2048 points
Source of FFT
Window
Analog channels 1 or 2 (or 3 or 4 on 54624A only), 1+2, 1-2, 1*2
Rectangular, Flattop, Hanning
-70 to -100 dB depending on averaging
In dBV, dBm@50 Ω
Noise Floor
Amplitude Display
Frequency Resolution
Maximum Frequency
0.097656/(time per div.)
102.4/(time per div.)
*
Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and 10 °C from firmware calibration temperature.
18
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Characteristics (continued)
Storage
Save/Recall (non-volatile)
Floppy Disk
54621A/D, 54622A/D, 54624A: 3 setups and traces can be saved and recalled internally
54641A/D, 54642A/D: 4 setups and traces can be saved and recalled internally
3.5” 1.44 MB double density
Image formats : TIF, BMP
Data formats: X and Y (time/voltage) values in CSV format
Trace/setup formats: Recalled
I/O
RS-232 (Serial) Standard Port
Parallel Standard Port
Printer Compatibility
1 port: XON or DTR; 8 data bits; 1 stop bit; parity = none; 9600, 19200, 38400, 57600 baud rates (use Agilent 34398A cable)
Printer support
HP DeskJet, LaserJet with HP PCL 3 or greater compatibility:
Black and white @150x150 dpi; Gray scale @ 600x600 dpi
Epson: black and white @180x180 dpi
Seiko thermal DPU-414: black and white
Optional GPIB Interface
Module (N2757A)
Fully programmable with IEEE488.2 compliance
Typical GPIB throughput of 20 measurements or twenty 2000-point records per second
Optional Printer Kit
The N2727A is a thermal printer kit, includingprinter power, parallel cable, power cable and paper
General Characteristics
Physical
Size: 32.26 cm wide x 17.27 cm high x 31.75 cm deep (without handle)
Weight: 6.35 kgs (14 lbs) on 54621A/D, 54622A/D, 54624A; 6.82 kgs (15 lbs) on 54641A/D, 54642A/D
Probe Comp Output
Trigger Out
54621A/D, 54642A/D, 54624A: Frequency ~1.2 kHz; Amplitude ~5 V
54641A/D, 54642A/D: Frequency ~2 kHz; Amplitude ~5 V
54621A/D, 54622A/D, 54624A: 0 to 5 V with 50 Ω source impedance; delay ~ 55 ns
54641A/D, 54642A/D: 0 to 5 V with 50 Ω source impedance; delay ~ 22 ns
Printer Power
7.2 to 9.2 V, 1 A
Kensington lock
Connection on rear panel for security
*
Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and 10 °C from firmware calibration temperature.
19
Download from Www.Somanuals.com. All Manuals Search And Download.
Performance Characteristics (continued)
Power Requirements
Line Voltage Range
54621A/D, 54622A/D, 54624A: 100 - 240 VAC 10%, CAT II, automatic selection
54641A/D, 54642A/D: 100-240 VAC, 50/60 Hz, CAT II, automatic selection; 100-132 VAC, 440 Hz, CAT II,
automatic selection
Line Frequency
Power Usage
54621A/D, 54622A/D, 54624A: 47 to 440 Hz
54641A/D, 54642A/D: 50/60 Hz, 100-240 VAC; 440 Hz, 100-132 VAC
54621A/D, 54622A/D, 54624A: 100 W max
54641A/D, 54642A/D: 110 W max
Environmental Characteristics
Ambient Temperature
Humidity
Operating -10 °C to +55 °C; Non-operating -51 °C to +71 °C
Operating 95% RH at 40 °C for 24 hr; Non-operating 90% RH at 65 °C for 24 hr
Operating to 4,570 m (15,000 ft); Non-operating to 15,244 m (50,000 ft)
Altitude
Vibration
HP/Agilent class B1 and MIL-PRF-28800F; Class 3 random
1
Shock
HP/Agilent class B1 and MIL-PRF-28800F; (operating 30 g, / sine, 11-ms duration, 3 shocks/axis along major axis.
2
Total of 18 shocks)
Pollution Degree2
Indoor Use Only
Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation must be expected.
This instrument is rated for indoor use only
Other
Installation Categories
CAT I:Mains isolated
CAT II:Line voltage in appliance and to wall outlet
Regulatory Information:
Safety
IEC 61010-1:1990+A1:1992+A2:1995 / EN 61010-1:1994+A2:1995
UL 3111
CSA-C22.2 No. 1010.1:1992
Supplementary Information
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive
89/336/EEC, and carries the CE-marking accordingly. The product was tested in a typical configuration with
HP/Agilent test systems.
*
Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and 10 °C from firmware calibration temperature.
20
Download from Www.Somanuals.com. All Manuals Search And Download.
Ordering Information
Description
54621A
54621D
54622A
54622D
54624A
54641A
54641D
54642A
54642D
2-channel 60 MHz Oscilloscope
2+16 channel 60 MHz Mixed Signal Oscilloscope
2-channel 100 MHz Oscilloscope
2+16 channel 100 MHz Mixed Signal Oscilloscope
4-channel 100 MHz Oscilloscope
2-channel 350 MHz Oscilloscope
2+16 channel 350 MHz Mixed Signal Oscilloscope
2-channel 500 MHz Oscilloscope
2+16 channel 500 MHz Mixed Signal Oscilloscope
Accessories Included: 54621A 54621D 54622A 54622D 54624A 54641A 54641D 54642A 54642D
User’s Guide (localized),
Service Manual,
√
√
√
√
√
√
√
√
√
Programmer’s Manual
Power cord
√
√
√
√
√
√
√
√
√
10073C 10:1 divider
probes with readout
0
0
0
0
0
2
2
2
2
10074C 10:1 divider
probes with readout
2
2
2
2
4
0
0
0
0
16:2 x 8 input logic
input probe assembly
√
√
√
√
√
√
√
√
√
√
Accessories pouch
and front panel cover
√
√
√
√
√
√
√
√
IntuiLink software
RS-232 cable
**
**
Manual Options (if no option is specified, English ABA will be shipped)
ABA
ABJ
AB2
English
Japanese
Simplified Chinese
21
Download from Www.Somanuals.com. All Manuals Search And Download.
Ordering Information (continued)
Available Options
Description
Option 003
Shielding Option for use in severe environments or with
sensitive devices under test – shields both ways (in and out):
RS-03 Magnetic interface shielding added to CRT, and RE-02
display shield added to CRT to reduce radiated interference
Option 1CM
Rackmount kit (same as 1186A)
.
Warranty and Calibration Options:
All models include a standard 1-year warranty. Contact local sales office for prices of extended options:
A6J
ANSI/NSCL Z540 calibration with test data (replaces 1BP)
3-years of return-to-Agilent inclusive calibration service
3-years of return-to-Agilent inclusive plus calibration service
5-years of return-to-Agilent inclusive calibration service
5-years of return-to-Agilent inclusive plus calibration service
1-year return-to-Agilent warranty, extended to 3-years
1-year return-to-Agilent warranty, extended to 5-years
R-50C-011-3
R-50C-013-3
R-50C-011-5
R-50C-013-5
R-51B-001-3C
R-51B-001-5C
Accessories
1183A
Description
Testmobile scope cart
Carrying case
1185A
1186A
Rackmount kit (same as option 1CM)
N2726A
Accessory pouch & front panel cover
(standard with 100-MHz models, optional with 60-MHz models)
N2727A
Seiko thermal printer and pouch, parallel cable,
power cable, 2 rolls paper, front panel cover
N2728A
N2757A
N2758A
10 rolls of printer paper
GPIB interface module for 54620-Series and 54640-Series scopes
CAN trigger module
Passive Probes
10070C
Description
1:1 passive probe with ID
10:1 500 MHz passive probe with ID
10:1 150 MHz passive probe with ID
10073C
10074C
22
Download from Www.Somanuals.com. All Manuals Search And Download.
Ordering Information (continued)
Fine Pitch Probing
10072A
Description
Fine-pitch probe kit
10075A
0.5 mm, IC clip kit
E2613B
0.5mm, Wedge probe adapter, 3-signal, qty 2
0.5 mm, Wedge probe adapter, 8-signal, qty 1
0.5 mm, Wedge probe adapter 16-signal, qty 1
0.65mm, Wedge probe adapter, 3-signal, qty 2
0.65 mm, Wedge probe adapter, 8-signal, qty 1
0.65 mm, Wedge probe adapter, 16-signal, qty 1
E2614A
E2643A
E2615B
E2616A
.
.
E2644A
Current Probes
1146A
Description
100 kHz current probe, ac/dc
50 MHz current probe, ac/dc
Power supply for N2774A
N2774A
N2775A
High Voltage Probes Description
10076A
N2771A
100:1, 4 kV 250 MHz probe with ID
1000:1, 15 kV, 50 MHz high voltage probe.
Logic Probes
Description
10085-68701
16:16 logic cable and terminator (for use with
54621D/622D/641D/642D)
54620-68701
16:2 x 8 logic input probe assembly (shipped standard
with 54621D/622D/641D/642D)
Active probes
1144A
Description
800 MHz active probe
1145A
2-channel 750MHz active probe
power supply for 1144A and 1145A
1142A
Differential Probes Description
N2772A
N2773A
20 MHz differential probe
Differential probe power supply
Cables
10833A
34398A
Description
GPIB cable, 1m long
RS-232 cable (standard with 100 MHz, 350 MHz,
and 500 MHz scopes)
23
Download from Www.Somanuals.com. All Manuals Search And Download.
By internet, phone, or fax, get assistance with
all your test & measurement needs
Call the measurement experts at Agilent Technologies
Whether your work is mostly digital, mostly analog, or somewhere in the middle, our
measurement specialists can help you select the best debugging solution. Call today to
talk to a knowledgeable engineer about your particular application.
Online assistance:
See for yourself with a no-obligation, 2-week demo
Phone or Fax
United States:
(tel) 800 829 4444
If you'd like to see for yourself how an Agilent 54600-Series scope can help you solve your
mixed analog and digital design problems, just give us a call. We'll arrange for you to use a
54620-Series scope in your lab for two weeks with no obligation to buy. Call 1-800-452-4844
and ask for Direct Demo or visit our website: www.agilent.com/find/MSO.
Canada:
(tel) 877 894 4414
(fax) 905 282 6495
Money-back guarantee
Get a full refund if you’re not satisfied with your purchase for any reason within 60 days in
the USA or 90 days outside of the USA.
China:
(tel) 800 810 0189
(fax) 800 820 2816
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and
problems. We strive to ensure that you get the test and measurement capabilities you paid
for and obtain the support you need. Our extensive support resources and services can help
you choose the right Agilent products for your applications and apply them successfully.
Every instrument and system we sell has a global warranty. Support is available for at least
five years beyond the production life of the product. Two concepts underlie Agilent's overall
support policy: "Our Promise" and "Your Advantage."
Europe:
(tel) (31 20) 547 2323
(fax) (31 20) 547 2390
Japan:
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
Korea:
(tel) (82 2) 2004 5004
(fax) (82 2) 2004 5115
Our Promise
Our Promise means your Agilent test and measurement equipment will meet its advertised
performance and functionality. When you are choosing new equipment, we will help you
with product information, including realistic performance specifications and practical
recommendations from experienced test engineers. When you use Agilent equipment, we
can verify that it works properly, help with product operation, and provide basic measurement
assistance for the use of specified capabilities, at no extra cost upon request. Many self-help
tools are available.
Latin America:
(tel) (305) 269 7500
(fax) (305) 269 7599
Taiwan:
(tel) 0800 047 866
(fax) 0800 286 331
Your Advantage
Your Advantage means that Agilent offers a wide range of additional expert test and
measurement services, which you can purchase according to your unique technical and
business needs. Solve problems efficiently and gain a competitive edge by contracting with
us for calibration, extra-cost upgrades, out-of-warranty repairs, and on-site education and
training, as well as design, system integration, project management, and other professional
engineering services. Experienced Agilent engineers and technicians worldwide can help
you maximize your productivity, optimize the return on investment of your Agilent
instruments and systems, and obtain dependable measurement accuracy for the life of
those products.
Other Asia Pacific Countries:
(tel) (65) 6375 8100
(fax) (65) 6836 0252
Email: [email protected]
Product specifications and descriptions in this
document subject to change without notice.
© Agilent Technologies, Inc. 2004
Printed in USA
May 1, 2004
5968-8152EN
Get the latest information on the products and applications you select.
Agilent T&M Software and Connectivity
Agilent's Test and Measurement software and connectivity products, solutions and
developer network allows you to take time out of connecting your instruments to your
computer with tools based on PC standards, so you can focus on your tasks, not on your
MICROSOFT and VISUAL STUDIO .NET are registered trademarks
of Microsoft Corporation in the U.S. and/or other countries.