SUPPLEMENT
3-447-108-27
3/1.24
Z502H / Z502O / Z505U – AKKUPACK
1
SAFETY INSTRUCTIONS
3
PRODUCT
3.1
SCOPE OF DELIVERY
Read and follow these instructions carefully and com-
pletely in order to ensure safe and proper use.
The instructions must be made available to all persons
who use the product.
1
1
Akkupack (Z502H / Z502O / Z505U)
Supplement (this document)
Keep for future reference.
3.2
SYMBOLS ON THE PRODUCT
Observe and comply with all safety regulations which are appli-
cable for your work environment.
Use the battery set in undamaged condition only.
Inspect the battery set before use. Pay particular attention to
damage.
Symbol Meaning
European conformity marking
Do not use the battery set after long periods of storage under
unfavorable conditions (e.g. humidity, dust or extreme tempera-
ture).
The product may not be disposed of with household
trash. You are required to comply with all applicable
The battery set must not be exposed to direct sunlight.
Only use the battery set in compliance with the specified tech-
nical data.
Batteries may only be charged in undamaged condition.
Inspect the batteries before use. Pay particular attention to
leaky and damaged batteries.
The battery set must be fed into the recycling loop for
rechargeable nickel metal hydrid batteries.
NiMH
3.3
TECHNICAL DATA
Unauthorized modification or opening of the accessory is pro-
hibited.
Z502H: 2000 mAh
Z502O: 2500 mAh
Z505U: 2620 mAh
Charging
capacity
2
APPLICATION
Nominal
voltage
Please read this important information!
9.6 V
2.1
INTENDED USE /
USE FOR INTENDED PURPOSE
Overall
Resistor
0.6 Ω (1000 Hz)
Z502H / Z502O / Z505U, hereinafter referred to as “product”, are
battery packs for supplying the following test instruments with elec-
trical power:
Operating temperatures: see test instrument
Storage temperatures: –20 °C … +30 °C
Ambient
Conditions
Storage duration: max. 1 year
Z502H:
all test instruments of the PROFITEST MASTER series
all test instruments the PROFITEST MF series
PROFITEST INTRO
Relative atmospheric humidity: 65 % 20 %
Terminals: for 9-V block battery clips
W × H × D: approx. 63 mm × 30 mm × 61 mm
Weight: approx. 250 g
Mechanical
Design
METRISO XTRA
Z502O: all test instruments the PROFITEST MF series
Z505U:
all test instruments of the PROFITEST MASTER series
all test instruments the PROFITEST MF series
PROFITEST INTRO
4
STORAGE AND TRANSPORT
METRISO XTRA
ATTENTION
Improper Storage
Safety of the user, as well as that of the product, is only assured
when it’s used for its intended purpose.
Damage to the product.
2.2
USE FOR OTHER THAN INTENDED PURPOSE
Store the product in a protected location and only within the limits
of the permissible ambient conditions (
Use of the product for any purposes other than those described in
this supplement, or in the product documentation for the associat-
ed test/measuring instrument, is contrary to use for intended pur-
pose. Use for Other than Intended Purpose may lead to
unpredictable damage!
ATTENTION
Improper Transport
Damage to the product.
2.3
LIABILITY AND GUARANTEE
Transport the product only within the limits of the permissible
Ensure adequate protection when transporting the product.
Liability and guarantee granted by Gossen Metrawatt GmbH com-
plies with the applicable contractual and mandatory legal regula-
tions.
1 | 2