Graham Field Photo Scanner 9000APS User Manual

INSTRUCTIONS FOR USE  
SERVICE MANUAL  
To avoid personal injury or damage to bed,  
please read all sections pertaining to  
your bed model before use.  
ETL/UL/CSA APPROVED  
UL 60601-1  
IEC 60601-2-38  
CSA 22.2 601.1  
Basic American Medical Products, a division of GF Health Products, Inc.  
Zenith APS Service Manual 999-0831-190A  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
TABLE OF CONTENTS  
Safety & Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4  
ID/Warning Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6  
Entrapment/Compliance . . . . . . . . . . . . . . . . . . . . . . . . . 7  
Features & Specifications . . . . . . . . . . . . . . . . . . . . . . . . 8  
Unpacking Your New Bed . . . . . . . . . . . . . . . . . . . . . . . . 9  
Assembly - Boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10  
Plugging in Your Staff Control . . . . . . . . . . . . . . . . . . . . 11  
Standard Mattress Retainers . . . . . . . . . . . . . . . . . . . . . 12  
Standard Wireform Wallsaver . . . . . . . . . . . . . . . . . . . . 13  
Wallsaver Adapter for Trendelenberg Positions . . . . . . .14  
Operations/Controllers & Staff Control . . . . . . . . . .15 - 16  
Placing the Bed into the Chair Position . . . . . . . . . . . . .17  
Using Trendelenberg/Reverse Trendelenberg . . . . . . . .18  
Electronic Components . . . . . . . . . . . . . . . . . . . . . . . . . 19  
Electrical Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . 20 - 22  
Optional Pivot Assist Bar . . . . . . . . . . . . . . . . . . . . . . . .23  
Optional Trapeze Support Adaptor . . . . . . . . . . . . . . . . .24  
Optional Battery Backup Configurations . . . . . . . . . . . . .25  
Troubleshooting - No Power . . . . . . . . . . . . . . . . . . . . . 26  
Troubleshooting - Staff Control & Pendant . . . . . . . 26 -27  
Troubleshooting Motors/Control Box . . . . . . . . . . . . . . . 28  
Replacing the Staff Control Assembly . . . . . . . . . . . . . . 29  
Maintenance & Inspection . . . . . . . . . . . . . . . . . . . . . . .30  
Replacement/Service Part List . . . . . . . . . . . . . . . . 31-32  
Service Part Diagrams . . . . . . . . . . . . . . . . . . . . . . . 33-36  
lABel syMBol deFInItIons  
!
!
Consult  
Accompanying  
Documents  
Safe  
Working  
Load  
Double  
Insulated  
Protected  
Grounded  
Device  
Type B  
Equipment &  
Applied Parts  
3
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
IMPORTANT SAFETY AND WARNING INFORMATION  
This product is a variable height, adjustable mattress  
platform, which will provide comfort and convenience  
for residents/patients and caregivers in long term  
care settings.  
The bed should ALWAYS be left in its lowest position  
when unattended to reduce the risk of injury while getting  
in or out of the bed.  
!
!
!
!
!
Keep all moving parts free of obstructions (i.e. blankets/  
sheets, heating blankets/pads, wiring, etc.)  
The MAXIMUM SAFE WORKING LOAD for the  
APS 9000 Bed with weight evenly distributed, including  
bedding, resident/patient, support surface, and all  
accessories, is 600 lbs.  
DO NOT use the assist devices as push handles for moving  
the bed. Assist devices can be deformed or broken if  
excessive side pressure is exerted. Assist devices are not  
meant for patients considered as high risks for entrapment  
(i.e. patients with pre-existing conditions such as  
confusion, restlessness, lack of muscle control, altered  
mental status, either organic or medicinal, or a combi-  
nation thereof). Additional safety measures should be  
considered for such high-risk patients.  
!
NEVER operate the bed if a Power Cord or Plug is  
damaged or not working properly. Contact qualified  
Service Personnel for examination and repair. Always  
unplug the Power Cord when performing any mainte-  
nance on the bed.  
DO NOT open assemblies such as the Actuators, Hand  
Control, or Control Box. If unauthorized personnel  
perform work on these components, the manufacturer’s  
warranty becomes void.  
NEVER permit more than one (1) person on/in the bed  
at any time.  
!
!
!
Body weight should be evenly distributed over the  
sleeping surface of the bed. DO NOT allow the patient  
to lay, sit, or lean in such a way that their entire body  
weight is placed only on the raised head or foot sections  
of the bed. This especially applies when repositioning  
or transferring a patient in or out of the bed. Increased  
risk may occur when the patient’s size and/or weight are  
inappropriate for the bed’s dimensions or weight capacity.  
DO NOT use unauthorized parts, accessories, or  
adaptors other than those specified/authorized by  
GF Health Products, Inc.  
!
!
When operating the High/Low, Knee, or Back Functions  
of the bed,ALWAYS ensure that the confined individual  
is positioned properly within the confines of the bed.  
DO NOT let any extremities protrude over the side or  
between the bed rails when performing these functions.  
The bed is intended for use, storage, and transport within  
a temperature range of -40˚C to +60˚C. It has a water-  
resistance rating of IPX4 and IS NOT to be power-  
washed or submersed.  
!
DO NOT lower the bed when objects are beneath it.  
Failure to inspect under the bed can result in damage  
to property or personal injury.  
!
!
operating Conditions  
Operation of the bed is based on the following conditions:  
Ambient Temperature of +10˚C to +40˚C; Relative Humidity  
of 30% to 75% (Non-condensing); Atmospheric Pressure of  
700hPa to 1060hPa; and a Splash Protection of IEC 60529.  
The bed’s Pendant Cord MUST BE ROUTED AND  
SECURED PROPERLY to ensure it does not become  
entangled and eventually severed during use. Also  
make sure all electrical cords DO NOT get tangled  
around the bed, side rails, or legs during transport or  
normal operation of the bed.  
storage  
Storage of the bed is based on the following conditions:  
Ambient Temperature of -10˚C to 70˚C; Relative Humidity  
of 10% to 100%; and an Atmospheric Pressure of 500hPa  
to 1060hPa.  
When using nasal- or masked-type administering  
equipment, all oxygen or air tubing MUST BE ROUTED  
AND SECURED PROPERLY to ensure that the tubing  
does not become entangled and eventually severed  
during the normal operation of the bed.  
!
Radio Frequency Interference (RFI)  
RFI influences most electronic equipment. Caution should be  
exercised with regard to the use of portable communications  
equipment in the area around such equipment. If RFI causes  
erratic behavior, shut the bed off immediately. Leave it off  
while the transmission is in progress.  
4
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
TYPICAL APS 9000 BED IDENTIFICATION LABELS  
with Grounded Electrical Cable  
Bed labels are an important part of  
MODEL NO  
identifying your bed’s make and model  
when ordering replacement parts. The  
Serial Number is essential if you are  
claiming parts or service under warranty.  
These helpful labels can be located on  
the main frame rails, immediately below  
the sleep decks on either side of the bed.  
GF HEALTH  
PRODUCTS, INC.  
806-0000-000000  
ETL  
Logo  
MFG. DATE  
12/23/11  
VOLT HZ AMP TYPE  
110-120 50/60 4.5 AC  
DUTY CYCLE - 2 min Continuous/18 min  
MODE OF OPERATION - Intermittent  
600 lbs  
=
SERIAL NO  
SAFE WORKING LOAD  
!
IPX4  
000000000000  
Please have this IMPORTANT infor-  
mation ready when calling our customer  
service or technical support staff at  
800-365-2338; it will allow us to better  
assist you and quickly answer your  
questions and concerns.  
SERIALAPS 9  
ZENITH 806 SERIES  
PATENT PENDING  
APS9000  
APS98174  
MODEL NO. 806-0000-000000  
MFG DATE 12/23/11  
000000000000  
SERIALAPS98174  
TYPICAL APS 9000 BED IDENTIFICATION LABELS  
for Class II Models  
If you opted for an electrical system without  
grounding on your electrical cord, your bed  
is designated as a Class II.  
MODEL NO  
806-0000-000000  
GF HEALTH  
PRODUCTS, INC.  
ETL  
Logo  
MFG. DATE  
12/23/11  
The Identification bed labels to the left  
represent the labels for Class II beds.  
They can be located on the main frame  
rails, immediately below the sleep decks  
on either side of the bed.  
VOLT HZ AMP TYPE  
110-120 50/60 4.5 AC  
DUTY CYCLE - 2 min Continuous/18 min  
MODE OF OPERATION - Intermittent  
600 lbs  
=
SERIAL NO  
SAFE WORKING LOAD  
!
IPX4  
000000000000  
Please have this IMPORTANT information  
ready when calling our customer service  
or technical support staff at 800-365-2338.  
The labels at left will allow us to better  
assist you in answering your questions  
and concerns.  
SERIALAPS 9  
ZENITH 806 SERIES  
PATENT PENDING  
APS9000  
APS98174  
MODEL NO. 806-0000-000000  
MFG DATE 12/23/11  
000000000000  
S
SERIALAPS98174  
5
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
TYPICAL SAFETY/WARNING BED LABELS  
The following warning labels have been placed on your bed to help protect you and your bed from  
damage. Please do not remove any labels from your bed.  
WARNING!  
CAUTION  
DO NOT LOWER BED WHEN OBJECTS ARE  
BENEATH BED. FAILURE TO INSPECT UNDER  
BED CAN RESULT IN DAMAGE TO PROPERTY  
OR PERSONAL INJURY.  
THIS BED IS SUITABLE FOR USE ONLY WITH  
OXYGEN ADMINISTERING EQUIPMENT OF  
THE NASAL OR MASK TYPE OR A TENT  
COVERING ONLY THE UPPER HALF (HEAD  
END) OF THE BED. OXYGEN TENT CANOPIES  
SHOULD NOT EXTEND BELOW BED SPRING  
LEVEL. LOCK HAND CONTROL AT FOOT OF  
BED WHEN USING OXYGEN ADMINISTERING  
EQUIPMENT.  
ATTENTION:  
S’assurer de ne pas faire descendre le lit  
lorsque des objets se trouvent sous le lit. Ne  
pas inspecter le dessous du lit pourrait entrainer  
des dommages materiels et des risques de  
blessures.  
ATTENTION:  
CE LIT PEUT ETRE UTILISE UNIQUEMENT  
AVEC UN EQUIPMENT DESTINE A  
L’ADMINISTRATION D’OXYGENE DE TYPE  
NASAL OU MASQUE OU AVEC UNE TENTE  
RECOUVRANT SEULEMENT LA MOTTIE  
AVENT (TETE) DU LIT. LES COTES DE LAS  
TENTE OXYGENE NE DOIVENT PAS SE  
PROLONGER PLUS DAS QUE LA SOMMIER  
DU LIT.  
WARNING!  
DO NOT LOWER BED WHEN  
OBJECTS ARE BENEATH BED.  
FAILURE TO INSPECT UNDER  
BED CAN RESULT IN DAMAGE  
TO PROPERTY OR PERSONAL  
INJURY.  
6
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
ENTRAPMENT & COMPLIANCE INFORMATION  
On March 10, 2006, the FDA (U.S. Food and Drug Adminis-  
tration) released long-awaited guidelines for reducing the risk  
of bed entrapment: “Hospital Bed System Dimensional and  
Assessment Guidance to Reduce Entrapment”. The new Guidance  
identifies potential entrapment areas and those body parts most  
at risk for entrapment; provides design criteria for manufacturers  
of new hospital/convalescent beds; recommends particular test  
methods to assess the conformance of existing hospital/conva-  
lescent bed systems; and answers frequently-asked questions  
about entrapment issues.  
It is also essential to have the correct bed components/acces-  
sories that correspond with the needs of your patient/resident  
and the particular bed you have purchased. Matching the correct  
bed component that correlates with the regulatory guidelines can  
be a daunting task. Our sales team at GF Health Products, Inc.  
and our friendly Customer Service Representatives at Basic  
American Medical Products can help you sift through the wide  
array of compliance and bed options. We will help you determine  
which bed/bed part is best for your patient’s/resident’s particular  
needs and help you with your compliance issues.  
The new Guidance was a result of a long-standing collaboration  
between the FDA and the Hospital Bed Safety Workgroup  
(HBSW), formed in 1999. GF Health Products, Inc’s Long-  
Term Care Bed division: Basic American Medical Products, is  
an HBSW charter member. As a result of our commitment to  
product safety, all our current long-term care beds have been  
strictly tested and conform to the new FdA Guidance.  
the APs 9000 bed and accessories listed in this  
manual are in full compliance with FdA guidelines  
for reducing the risk of bed entrapment: “Hospital  
Bed System Dimensional and Assessment Guidance  
to Reduce Entrapment”.  
The guidelines set forth by the FDA Guidance layout specific  
dimensional limitations on potentially injury-threatening gaps  
and spaces that can occur between bed system components,  
such as rails, when not properly installed. GF Health Products,  
Inc. and Basic American Medical Products have conformed  
to these guidelines from a manufacturing aspect. However,  
entrapment issues can often arise when a healthcare provider/  
facility has not correctly assembled the components on a bed.  
It is essential that the provider/facility fully understand their  
responsibility in complying to the guidelines set forth by the  
FDA in order to avoid injury. To that end, we have provided  
the FDA’s web address at right as a resource for understanding  
and following these guidelines for the safety of patients/residents.  
7
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
ZENITH 9000 APS MECHANICS  
NOTE: All dimensions are in a range of +/- .25 inches  
Overall Bed Length (with brds & wallsaver) . . 82”/86”  
Overall Bed Width (with boards) . . . . . . . . . . . 36.00”  
High Height* . . . . . . . . . . . . . . . . . . . . . . . . . . 30.00”  
Low Height* . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.00”  
Maximum Head/Back Deck Angle . . . . . . . . . . . . 70°  
Maximum Knee/Foot Deck Angle . . . . . . . . . . . . 30°  
Maximum Safe Working Load . . . . . . . . . . . 600 lbs.  
WITH WEIGHT EVENLY DISTRIBUTED  
- includes bedding, resident, support surface,  
and all accessories.  
Mass of bed (without assist devices or boards) = 204 lbs.  
* Bed height calculated from floor surface to top of  
sleep deck.  
ZENITH 9000 APS ELECTRICAL  
Power/Frequency . . . . . . . . . . . . . 120 Volt 50/60 Hz  
Output Rating . . . . . . . . . . . . . . . . . . . . 12/33V IPX4  
Overall Movement Draw . . . . . . . . . . . . . 4.50 Amps  
Classification . . . . . . . . . . . . . . . . . . . Class I, Type B  
Electrical Cord . . . . . . . . . . . . #18 AWG 3 Conductor  
Type SJT  
Classification . . . . . . . . . . . . . . . . . . Class II, Type B  
Electrical Cord . . . . . . . . . . . . #18 AWG 2 Conductor  
Type SJT  
Mode of Operations . . . . . . . . .10% Max. Duty Cycle  
2 minutes on/18 minutes off  
Battery Backup Option 1 . . . . 120V External Battery  
and Charger  
Pack and Charger can be purchased separately  
as an accessory.  
Battery Backup Option 2 . . . . 120V External Battery  
Pack is mounted to the bed frame -  
no charger needed.)  
8
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
DISCARD  
KEEP  
UNPACKING YOUR BED  
1. Cardboard Blocks  
2. Large Block  
9. Wireform Wallsaver  
10. Two Mattress Retainers  
11. Service Manual/Documents  
12. Wallsaver Adaptor Brackets for  
Trendelenberg and Reverse  
Trendelenberg Positioning  
13. Pendant Holster  
Make sure all parts/components are included.  
3. Large Cable Tie - CUT  
4. Notched 2 x 4 Board  
5. Foam Wrapping  
6. End Caps with Foam  
7. Notched Leg Foam  
8. Small Cable Tie - CUT  
Check all bed components for obvious damage.  
Inspect the Power Supply Cord for any cuts  
and/or damage.  
Check to see all actuator/motor cables are routed  
and connected properly to the control box.  
(attached to pendant)  
NOTE: END OF POWER CABLE  
IS COILED FOR SHIPPING AND  
TIED, WITH CABLE TIE, TO GRID  
WIRE ALONG WITH PENDANT,  
PENDANT CABLE & HOLSTER.  
2
9
4
PLEASE CUT AND DISCARD  
CABLE TIE AROUND CABLES  
WHEN YOU UN-PACK YOUR BED.  
3
6
8
1
5
10  
6
10  
NOTE:  
Underbed Light is optional and available on  
APS981741 Models.  
Underbed Light and Mounted Battery optional  
combination is available on APS981742 Models.  
13  
7
Cut and  
discard  
cable tie.  
12  
11  
7
ALSO  
DISCARD TOP  
AND BOTTOM  
CARTONS  
9
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
HEAD- AND FOOTBOARD ASSEMBLY  
2. FOOTBOARD INSTALLATION  
1. HEADBOARD INSTALLATION  
• The headboard comes with four pre-installed  
inserts - consider this the inside of the board.  
• The footboard comes with four pre-installed  
inserts - consider this the outside of the board.  
• Position 2 mounting tubes on the outside of  
the headboard with “L” facing inward.  
• Position 2 mounting tubes on the inside of the  
footboard with “L” facing outward.  
• Align the top hole of the mounting tubes with  
the top holes in the headboard.  
• Align the top hole of the mounting tubes with  
the top holes in the footboard.  
• Insert a 40mm hex drive bolt through each of  
the top holes and bottom holes and screw into  
each insert. Tighten with the Hex Allen wrench  
included in your kit.  
• Insert a 40mm hex drive bolt through each of  
the top holes and bottom holes and screw into  
each insert. Tighten with the Hex Allen wrench  
included in your kit.  
• Slide the “L” portions of the Mounting tubes  
into the hollow ends of the main frame rails,  
at the head deck end.  
• Slide the “L” portions of the Mounting tubes  
into the hollow ends of the main frame rails,  
at the foot deck end.  
FOR 80” BEDS: Slide the mounting tubes in  
until the 1ST hole in the tubes lines up with  
the 1ST hole in the rails. See sample below.  
FOR 80” BEDS: Slide the mounting tubes in  
until the 1ST hole in the tubes lines up with  
the 2ND hole in the rails. See sample below.  
FOR 76” BEDS: Slide the mounting tubes in  
until the 1ST hole in the tubes lines up with  
the 2ND hole in the rails. See sample below.  
FOR 76” BEDS: Slide the mounting tubes in  
until the 1ST hole in the tubes lines up with  
the 3RD hole in the rails. See sample below.  
Inserts  
on outside  
HEADBOARD  
FOOTBOARD  
CENTER  
OF BED  
40mm  
Bolts  
Inserts  
on inside  
CENTER  
OF BED  
40mm  
Bolts  
1st  
hole  
80” hole position  
(2nd hole in)  
1st  
hole  
76” hole  
76” hole position (3rd hole in)  
position  
80” hole  
(2nd hole in)  
position  
NOTE: The first hole at the foot end is reserved  
for attaching the Optional 4” pan extension.  
(1st hole in)  
10  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
PLUGGING IN THE FOOTBOARD NURSE/STAFF CONTROL  
STEP 1 - ATTACHING THE FOOTBOARD  
The APS 9000 bed features a Nurse/Staff Controller in the footboard, however the board is ordered separately with your  
bed because of the variety of board styles available. If ordered at the same time as the bed, the Staff Control Assembly  
and Shroud Cover will be pre-installed to the Footboard at the manufacturing factory.  
STEP 2 - CONNECTING YOUR CABLES  
Please refer to DETAIL A shown below.  
3KLOOLSVꢀ6FUHZV  
DWWDFKꢀ6KURXG  
ꢀ6WDIIꢀ&RQWURO  
$VVHPEO\ꢀIURP  
LQVLGHꢀRIꢀERDUG  
6WDIIꢀ&RQWURO  
$VVHPEO\  
a. If the Staff Control Assembly is not installed at the  
factory you will need to first attach it to the footboard.  
b. With the cutout side of the footboard facing outward  
as shown, insert the staff control cable through the  
large round hole.  
)227  
(1'  
c. With text right-side-up, insert the staff control  
assembly into the cutout slot on the footboard  
and attach to the board using the two outside  
small holes and two screws from your staff  
control hardware kit.  
6KURXG  
ꢁ&DEOHꢀ&RYHUꢂ  
&DEOHꢀ7LH  
d. Attach the Shroud Cover over the cable on the  
inside of the board using the four remaining  
screws from your kit.  
6WDIIꢀ&RQWURO  
$VVHPEO\  
FDEOHꢀWRꢀ  
PRXQWLQJꢀWXEH  
e. Insert the T-Cable end (extending out the  
foot end with phone jack) into the round  
plug, making sure the phone jack is  
seated correctly inside the female plug  
(arrow to arrow - see DETAIL B.)  
)RRWꢀ%RDUG  
ZLWKꢀFXWRXW  
f. Screw on the round lock cap onto  
the Staff control’s female plug to  
secure (see DETAIL B).  
SEE PAGE 10 OF THIS  
MANUAL FOR PROPER  
ASSEMBLY OF ELBOW  
MOUNTING BRACKETS  
(TUBES) & FOOTBOARD  
Loop Male Staff Control Cable  
and cable tie through mounting  
tube hole with Female Staff  
C
o
n
tr  
o
l
C
a
b
le
.
g. IMPORTANT:  
Make sure to tie  
$OLJQ  
$UURZV  
off the staff control  
cable to the foot-  
board mounting  
tube with a cable  
tie as shown.  
)HPDOHꢀHQGꢀRIꢀ6WDIIꢀ&RQWURO  
$VVHPEO\ꢀ&DEOHꢀꢃꢀSOXJVꢀLQWRꢀ7ꢃꢀ&DEOH  
/RFNLQJꢀ(QGꢀ&DSꢀRQꢀ'RXEOHꢀ7ꢃ&DEOH  
DETAIL B:  
MAKE SURE LOCK  
END CAPS ARE  
SCREWED ON  
SECURELY  
(QGꢀRIꢀ'RXEOHꢀ7ꢃ&DEOHꢀDWꢀIRRWꢀHQGꢀꢃꢀ*UH\  
ꢁꢄ7ꢄꢀUXQVꢀRQꢀHLWKHUꢀVLGHꢀRIꢀ6HDWꢀ3DQ  
IRUꢀDWWDFKLQJꢀ\RXUꢀ+DQGꢀ&RQWUROOHUꢅꢂ  
DETAIL A:  
CLOSEUP OF STAFF  
CONTROL CABLE CONNECTION  
11  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
STANDARD MATTRESS RETAINER INSTALLATION  
STEP 1. HEAD RETAINER INSTALLATION  
1. Position retainer with curved end toward headboard.  
2. Squeeze ends of the retainer toward the inside of  
the bed and lower between the deck rails and first  
outside grid wires.  
3. Align the ends of retainer with small holes on the  
inside of deck rails and slowly release.  
STEP 2. FOOT-END RETAINER  
INSTALLATION  
1. Position retainer with curved end toward footboard.  
2. Squeeze ends of the retainer toward the inside of  
the bed and lower between the deck rails and first  
outside grid wires.  
3. Align the ends of retainer with small holes on the  
inside of deck rails and slowly release.  
STEP 3. RETAINER POSITIONS  
FOR 80” BEDS: Use outside small holes.  
See sample at right.  
FOR 76” BEDS: Use inside small holes.  
See sample at right.  
Be sure to use a mattress that is properly sized to fit  
the sleep deck, which will remain centered on the  
deck relative to State and Federal Guidelines. Use  
of an improperly fitted mattress could result in injury  
or death.  
Use a properly sized mattress in order to minimize  
the gap between the side of the mattress and assist  
devices. This gap must be small enough to prevent  
resident/patient from getting his/her head or neck  
caught in this location.  
!
!
12  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
STANDARD WIREFORM WALLSAVER INSTALLATION  
STEP 1. WALLSAVER ASSEMBLY  
1. Position the Wireform Wallsaver (Part 999-0844-180) with bent  
end facing upward and tab ends facing inward as shown at right.  
2. Determine the position desired (See bottom of page).  
HEAD END  
3. Gently squeeze the tab ends of the wallsaver inward toward the  
center of the wallsaver and, holding the tabs parallel with the slots  
in the caster bases (See DETAILA), slide the tabs into the slots  
while letting the wallsaver gently expand outward.  
4. Turn the wallsaver downward until it rests on the floor (See  
Detail B).  
:LUHIRUPꢀ:DOOVDYHU  
ꢁꢁꢁꢂꢃꢄꢅꢅꢂꢆꢄꢃ  
%HQGꢀIDFHVꢀXSZDUG  
STEP 2. WALLSAVER  
REMOVAL  
1. To remove or move the Wireform  
Wallsaver to a new position, raise  
the wallsaver off the floor until  
the end tabs are horizontal.  
Finish with  
2. Squeeze the ends toward the  
center of the wallsaver until the  
end tabs slide out of the caster  
base slots.  
Wallsaver  
resting on  
ꢀ6WDUWꢀZLWKꢀWDEꢀKHOG  
ꢀKRUL]RQWDOO\  
floor  
STEP 2. WALLSAVER  
POSITIONS  
76” Bed Wallsaver  
Position w/ Trapeze or  
Position w/o Trapeze 80” Bed w/o Trapeze  
76” Bed Wallsaver  
80” Bed Wallsaver  
Position w/ Trapeze  
1. The APS 9000 wallsaver features  
3 positions for easy reconfiguration  
- 76”, 80”, and Trapeze (if you use  
Trendelenberg/Reverse Trendelen-  
berg positions please see optional  
adaptor brackets on next page).  
2. For 76” beds, position the wallsaver  
ends in the caster base slots closest  
to the FOOT end of the bed. For 80”  
beds, position the wallsaver ends  
in the MIDDLE slots on the bases.  
3. If you have an optional Trapeze unit  
on a 76” bed, position the wallsaver  
ends in the MIDDLE slots on the  
caster bases. For 80” beds position  
wallsaver in slots closest to HEAD.  
13  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
OPTIONAL WALLSAVER ADAPTOR BRACKET INSTALLATION TO  
USE WITH TRENDELENBERG/REVERSE TRENDELENBERG FEATURE  
3
4. Secure brackets to caster bases  
STEP 1. BRACKET  
2
using short Nylon Lock Nuts (#3  
INSTALLATION  
- 100-6744-002) included in your  
1. Remove the tall Lock Nut (#1)  
from the head end of both the  
right and left-hand caster bases.  
adaptor bracket bag/kit.  
3
2
2. Position the two saddle  
adaptor brackets (#2  
- 999-0831-002) so  
that the ends with  
slots face away from  
the bases as shown.  
1
3. Lower the brackets  
onto the caster bases,  
aligning holes in brackets  
with caster stems.  
NOTE: MOUNTING OF THE ADAPTOR  
BRACKETS IS ONLY NECESSARY IF YOU  
PLAN ON USING THE TRENDELENBERG/  
REVERSE TRENDELENBERG FEATURE.  
1
STEP 2. WALLSAVER  
ASSEMBLY  
1. Position the Wireform Wallsaver  
(Part 999-0844-180) with bent  
end facing upward and tab ends  
facing inward as shown at right.  
2. Gently squeeze the tab ends of  
the wallsaver inward toward the  
center of the wallsaver and, hold  
-ing the tabs parallel with the  
slots in the caster bases (See  
DETAILA), slide the tabs into the  
Finish with  
Wallsaver  
ꢀ6WDUWꢀZLWK  
ꢀWDEꢀKHOG  
ꢀKRUL]RQWDOO\  
resting on  
floor  
slots while letting the wallsaver gently expand outward.  
4. Turn the wallsaver downward until it rests on the floor (See Detail B).  
STEP 3. WALLSAVER REMOVAL  
1. To remove or move the Wireform Wallsaver to a new position, raise  
the wallsaver off the floor until the end tabs are horizontal.  
2. Squeeze the ends toward the center of the wallsaver until the  
end tabs slide out of the caster base slots.  
IMPORTANT:  
WALLSAVER END NEAREST THE WALL MUST FACE UPWARD  
AS SHOWN FOR WALLSAVER TO SEAT PROPERLY IN SLOTS.  
:LUHIRUPꢀ:DOOVDYHU  
ꢁꢁꢁꢂꢃꢄꢅꢅꢂꢆꢄꢃ  
%HQGꢀIDFHVꢀXSZDUG  
14  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
BED OPERATIONS - APS PENDANT/HAND CONTROLLER  
BACKLIT PENDANT  
1
2
3
4
5
6
7
8
9
HEAD DECK ANGLE UP BUTTON  
HEAD DECK ANGLE DOWN BUTTON  
HI/LO UP BUTTON (RAISE ENTIRE BED)  
HI/LO DOWN BUTTON (LOWER ENTIRE BED)  
KNEE & FOOT DECK ANGLE UP BUTTON  
KNEE & FOOT DECK ANGLE DOWN BUTTON  
SET TO CHAIR POSITION BUTTON  
UNDO CHAIR POSITION BUTTON  
OPTIONAL UNDERBED LIGHT BUTTON  
NON-BACKLIT PENDANT  
SAME FEATURES BUT WITHOUT  
UNDERBED LIGHT BUTTON  
OPERATION:  
1. To angle the Head Deck upward, PRESS 1  
2. To angle the Head Deck downward, PRESS 2  
3. To raise the Bed Up horizontally, PRESS 3  
4. To lower the Bed Down horizontally, PRESS 4  
5. To angle the Knee/Foot Decks upward, PRESS 5  
6. To angle the Knee/Foot Decks downward, PRESS 6  
7. To put the Bed in the Chair position, PRESS 7  
8. To release the Bed from the Chair position, PRESS 8  
9. To turn the underbed light on or off (*optional pendant only)  
PRESS THE LIGHT BULB ICON near the bottom of the  
hand controller (not shown on sample at left).  
Back of both  
Pendants  
include  
attachment  
hook.  
9
BACKLIT PENDANT  
Model #00HB851D-003  
Part # 999-0806-301  
Backlit and includes  
Underbed Light Button  
(button #9)  
Non-Backlit Pendant  
A Pendant  
holster also  
comes  
standard  
with the  
Model #00HB841A-001  
Part # 999-0806-305  
(without Underbed  
Light Button)  
APS bed.  
NOTE:  
THE ZENITH 9000 APS BED COMES STANDARD WITH A FOOTBOARD NURSE/STAFF CONTROL  
(PLEASE MAKE SURE TO ORDER WITH APPROPRIATE BOARD STYLE). THE BACKLIT PENDANT  
WITH UNDERBED LIGHT BUTTON AND THE UNDERBED LIGHT CAN BE PURCHASED AS OPTIONS..  
PLEASE SEE NEXT PAGE FOR NURSE/STAFF CONTROL OPERATING INSTRUCTIONS.  
15  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
BED OPERATIONS - NURSE/STAFF CONTROL PANEL  
1
ALL LOCKED BUTTON  
HEAD DECK UP BUTTON  
2
3
4
HEAD DECK DOWN BUTTON  
KNEE & FOOT DECK UP BUTTON  
KNEE & FOOT DECK DOWN BUTTON  
CHAIR POSITION BUTTON  
CPR/UNDO CHAIR POSITION BUTTON  
KEY LOCK/UNLOCK BUTTON  
HI/LO (T/TR) LOCKED BUTTON  
HI/LO UP BUTTON  
5
6
7
8
9
Press “Underbed Light” button (#14) to turn underbed light on or off.  
10  
11  
12  
13  
14  
15  
HI/LO DOWN BUTTON  
Press “CPR Heart” button (#7) to Level/Flatten the bed horizontally for CPR position.  
TRENDELENBERG  
“Safe Mode” (#15) LED light will glow green automatically once power cord is plugged in.  
Green light will blink intermittently for approximately 10 minutes. After 10 minutes of non-  
activity Hi/Lo Lock icons on the nurse/staff control panel will show orange (locked).  
REVERSE TRENDELENBERG  
UNDERBED LIGHT BUTTON  
SAFE MODE RELEASE BUTTON  
LOCK OUT SINGLE FUNCTIONS  
To individually lock out the “Head”, “Knee”, “Chair”,  
and “Hi/Lo” functions, press the appropriate top  
icon (#2, 3, 4, 10, or 12) button and the “Key”  
button at the same time. An orange LED lock  
symbol will appear under the related icon.  
LOCK SYMBOLS  
LOCK SYMBOL  
LOCK SYMBOLS  
To Unlock” any of the individual functions, press  
the top icon (#2, 3, 4, 10, or 12) and the “Key” Button  
(#8) simultaneously. LED lights will not show up.  
Press “Head”, “Knee”, or “Chair” and “Key” at the same time to unlock individual  
lock out functions.  
Press “Hi/Lo Locked” to lock out “Hi/Lo”, “Trendelenberg”,  
and “Reverse Trendelenberg” functions.  
LOCK OUT HI/LO FUNCTIONS  
To lockout the functions for raising and lowering  
the entire bed and tilting the bed for Trendelenberg  
positions, press the “Hi/Lo Locked” (#9) button. Orange  
LED lock symbols will appear under the icons.  
To Unlock” all Hi/Lo functions, press the “Hi/Lo  
Locked” button (#9) and the “Key” button (#8)  
simultaneously. Orange LED lights will not show up.  
Press “Hi/Lo Locked” and “Key” at the same time  
LOCK SYMBOLS  
to unlock all Hi/Lo lock out functions.  
LOCK OUT EVERYTHING  
LOCK SYMBOLS  
LOCK SYMBOL  
LOCK SYMBOLS  
To lock out all functions, press the “All Locked”  
button (#1). Orange LED lock symbols will appear  
under the “Head”, “Knee”, “Chair”, “Hi/Lo”, and  
“Trendelenberg/Reverse Trendelenberg” icons.  
To Unlock” all functions, press the “All Locked”  
button (#1) and the “Key” button (#8) simultaneously.  
Orange LED lights will not show up.  
Press “All Locked” and “Key” at  
the same time to unlock all functions.  
SAFE MODE:  
Green light will show automatically when bed is  
plugged in and will blink intermittently during 10  
minute “sleep mode”. After 10 minutes of lapsed  
time since the last function was used, the LED light  
shows green and Hi/Lo Lock icons will show orange.  
To manually disable the function, hold the “Safe  
Mode” button (#15) and “Key” button (#8) at the  
same time. Green LED light will go out.  
The LED will be lit green when Safe Mode automatically engages  
when plugging in the bed. Press “Safe Mode” and “Key” to disable  
the Safe Mode feature.  
16  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
BED OPERATIONS - CHAIR POSITION  
2
3
1
1. Make sure the Chair position on the staff  
control panel is not locked out (orange  
lock icon). To unlock, press the Key icon #1  
and Chair icon #2) (or All Locked button  
#3) at the same time.  
4
2. To move the bed to the Chair position  
(foot & knee decks angled up, head deck  
angled up, and head of bed tilted upward),  
press on the chair icon button (#2).  
HEAD  
END  
To release the chair feature and return  
the bed back into a horizontal position,  
press the CPR icon (#4). The Chair feature  
can also be operated using the Pendant.  
(See page 15.)  
FOOT  
END  
BED OPERATIONS - CPR/HEART POSITION  
3
1
2
1. Make sure the bed positions on the staff  
control panel are not locked out (orange  
lock icon). To unlock, press the Key icon (#1)  
and the All Locked button (#3) at the same  
time.  
2. To smoothly move the bed into a prone,  
horizontal position for CPR access, press  
and hold the CPR/Heart icon button (#2).  
CPR POSITION  
(The CPR button places the bed in a horizontal position)  
This illustration shows bed all the way down as a reference only -  
the CPR button will not lower the bed.)  
17  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
BED OPERATIONS - TRENDELENBERG/REVERSE TRENDELENBERG  
1
2
3
1. Make sure the Hi/Lo function on the staff  
control panel is not locked out (orange  
lock icon). To unlock, press the Key icon #1  
and Hi/LO Locked button #2 at the same  
time.  
2. To move the bed to the Trendelenberg  
position (foot end up), press on button #3.  
4
3. To move the APS bed into the Reverse  
Trendelenberg position (head end up),  
press on button #4.  
)227  
(1'  
75(1'(/(1%(5*  
+($'  
(1'  
NOTE: THE TRENDELENBERG AND REVERSE  
TRENDELENBERG FEATURE CAN ONLY BE  
ACCESSED USING THE STAFF CONTROL. IT  
CANNOT BE ACCESSED USING THE HAND  
CONTROLLER/PENDANT.  
+($'  
(1'  
5(9(56(ꢀ75(1'(/(1%(5*  
)227  
(1'  
18  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
APS 9000 SERIES ELECTRICAL COMPONENTS (LINAK)  
Control/Junction Box  
Part # 999-0831-300  
Quantity = 1  
STANDARD:  
Pendant/Hand Controller  
with Underbed Light Button  
Part # 999-0806-301  
Quantity = 1  
Power Cable plugs  
into side port (part  
listed at bottom)  
Bed also includes  
Pendant Holster  
Part # 999-0791-000  
Quantity = 1  
Head Motor/Actuator  
Part # 999-0822-052  
Quantity = 1  
Foot Motor/Actuator  
Part # 999-0822-053  
Quantity = 1  
Double “T” Cable  
Part # 999-0806-200  
Quantity = 1  
Motor Cable (Hd & Hi/Lo)  
Part # 999-0806-203  
Quantity = 3  
Motor Cable (Ft.)  
Part # 999-0806-202  
Quantity = 1  
Hi/Lo Motor/Actuator  
Part # 999-0831-051  
Quantity = 2  
Battery Backup Cable  
Part # 999-0831-200  
Quantity = 1  
STANDARD:  
3 Prong Power Cable  
Part # 999-0775-208  
Quantity = 1  
OPTIONAL UNDERBED LIGHT  
OPTIONAL:  
2 Prong Power Cable  
Part # 999-0775-206  
Quantity = 1  
Underbed Light Adaptor Cable  
Part # 999-0806-212  
Quantity = 1  
Underbed Light  
Part # 999-0806-220  
Quantity = 1  
DO NOT use unauthorized parts, accessories, or adaptors other than those specified/authorized  
by GF Health Products, Inc. DO NOT open assemblies such as the Actuators, Hand Control, or  
Control Box. If unauthorized personnel perform work on these components, the manufacturer’s  
warranty becomes void. NEVER operate the bed if a Power Cord or Plug is damaged or not working  
properly. Contact qualified Service Personnel for examination and repair. Always unplug the Power  
Cord when performing any maintenance on the bed.  
!
19  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
APS9000 SERIES ELECTRICAL CABLING - STANDARD  
u
v
w
x
y
z
{
|
POWER CABLE  
T-CABLE  
HEAD CABLE  
u
FOOT CABLE  
HI/LO CABLE (FOOT)  
HI/LO CABLE (HEAD)  
BATTERY BACKUP CABLE  
GROUND WIRE  
u
Portable  
Battery  
Cable  
Extension  
x
|
z
v
v
Ground  
Wire  
w
y
{
u
{
T-Cable (HB)  
Hi/Lo (Ft) - Port 4  
Foot - Port 3  
Hi/Lo (Hd) - Port 2  
Head - Port 1  
v
Power Cable &  
Battery Cable -  
Side of Control Box  
20  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
APS9000 SERIES ELECTRICAL CABLING - WITH UNDERBED LIGHT  
NOTE: Cabling for Beds with Underbed Light feature  
is identical to Standard APS beds, except that the  
T-Cable plugs into Port 2 in the Underbed Light Box  
and an extra Extension Cable is plugged into Port 1  
on the Underbed Light Box and the HB Port in the  
Control Box.  
u
v
w
x
y
z
{
POWER CABLE  
T-CABLE  
HEAD CABLE  
FOOT CABLE  
HI/LO CABLE (FOOT)  
HI/LO CABLE (HEAD)  
BATTERY BACKUP CABLE  
GROUND WIRE  
u
v
v
|
x
|
}
UNDERBED LIGHT CABLE (UBL)  
z
w
Portable  
Battery  
Cable  
y
Extension  
{
v
}
Ground  
Wire  
}
v {  
Light  
T-Cable - Port 2 in Underbed Light Box  
Hi/Lo (Ft) - Port 4 in Control Box  
Foot - Port 3 in Control Box  
Hi/Lo (Hd) - Port 2 in Control Box  
Head - Port 1 in Control Box  
Power Cable &  
Battery Cable -  
Side of Control Box  
UNDERBED  
LIGHT (UBL)  
21  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
APS9000 SERIES ELECTRICAL CABLING - WITH  
UNDERBED LIGHT & BATTERY  
u
| z y  
{
x
u
POWER CABLE  
w
v
T-CABLE  
w
HEAD CABLE  
x
FOOT CABLE  
y
HI/LO CABLE (FOOT)  
v
z
HI/LO CABLE (HEAD)  
}
{
BATTERY BACKUP CABLE  
|
GROUND WIRE  
}
UNDERBED LIGHT CABLE (UBL)  
{
Mounted  
Battery  
Cable  
v
Ground  
Wire  
{
Underbed Light (UBL)  
Attached  
Battery  
Cable  
NOTE: Cabling for Beds with Underbed Light and Mounted  
Battery Backup features are identical to Standard APS beds,  
except that the T-Cable plugs into Port 2 in the Underbed  
Light Box and an extra Extension Cable is plugged into Port 1  
on the Underbed Light Box and the HB Port in the Control Box.  
MOUNTED  
BATTERY  
OPTION  
A Battery Cable runs from the Control Box directly to an  
external Battery Pack at the foot end.  
22  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
OPTIONAL PIVOT ASSIST BAR INSTALLATION & OPERATION  
INSTALLATION  
1. At the head end of the bed, position the Pivoting Assist Bar (#2)  
bracket over the Head Deck (#1) frame and align the two holes  
in the Assist Bar bracket with the two holes in the frame rail on  
either side of the head deck.  
Position the Assist Bar  
Bracket ON TOP of the  
Deck Frame, aligning  
holes in the bracket with  
holes in the frame.  
2
2. From the outside, insert the Clevis Pins (#3) through the  
holes in the pivot bracket and deck frame.  
1
3
+($'  
(1'  
3
3. From the inside, insert a Hair Pin Clip (#4) through  
each of the small holes in the clevis pins.  
PLEASE ORDER KIT #ZA85400. KIT INCLUDES ONE PIVOTING  
ASSIST BAR ASSEMBLY AND TWO LANYARD ASSEMBLIES  
WITH CLEVIS PINS AND HAIR PIN CLIPS. (A STATIONARY ASSIST  
BAR IS ALSO AVAILABLE AS AN OPTION - KIT # ZA85000).  
4
4
THE ASSIST BAR CAN BE POSITIONED ON EITHER  
THE RIGHT SIDE OR LEFT SIDE OF THE HEAD DECK.  
IT IS NOT OFFERED FOR THE FOOT END OF THE BED.  
SAMPLE SHOWS PIVOTING  
ASSIST BAR POSITIONED ON  
THE LEFT SIDE OF THE BED.  
1. To release the pivoting assist bar from its  
vertical lock position, hold onto the top of  
the bar with one hand and slightly pull out  
the black knob on the outside of the pivot  
assembly with the other hand to release  
the locking mechanism. For ease of patient  
access, it is recommended that you pivot  
the rail toward the headboard until it stops  
and rests in place.  
PIVOTING ASSIST BAR OPERATION  
+($'  
(1'  
5HFRPPHQG  
3LYRWꢀ7RZDUG  
+HDGERDUG  
2. To set the assist bar back to its upright  
locked position, grab the top of the assist  
bar with one hand and pivot the assembly  
upward until the black knob mechanism  
snaps into place, locking the assist bar  
vertically.  
%ODFNꢀ.QREꢀWR  
/RFNꢀRUꢀ5HOHDVH  
ꢀꢆꢂꢃꢃꢄ  
ꢀꢁꢂꢃꢃꢄ  
ꢇꢂꢃꢃꢄ  
+($'  
'(&.  
(1'  
+($'ꢅ'(&.  
6/((3ꢅ685)$&(  
+($'  
'(&.  
(1'  
FOR YOUR SAFETY, MAKE SURE YOUR FINGERS  
ARE CLEAR OF THE SIDES OR UNDERNEATH THE  
PIVOTING ASSEMBLY WHEN PIVOTING THE  
ASSIST BAR UP OR DOWN.  
PLEASE MAKE SURE THE PIVOT ASSIST BAR  
IS ALWAYS IN THE FULL, VERTICAL, LOCKED  
POSITION WHEN RESIDENT/PATIENT IS LEFT  
UNATTENDED.  
!
!
23  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
OPTIONAL TRAPEZE SUPPORT INSTALLATION  
(The Trapeze Support is attached to the head section of the bed.)  
Please Order Kit Number ZA79000  
Trapeze Support vertical  
support tube. (TMI Bar  
Sleeve would be inserted  
inside the tube.)  
+($'  
(1'  
0DLQꢆ)UDPH  
Use first holes in  
Trapeze Support  
to mount to main  
frame tubes.  
USE SAME CLEVIS PINS AND HAIR PIN CLIPS  
Use third hole  
from the end of the  
main frame tubes.  
THAT WERE USED TO CONNECT HEAD BOARD  
TO MOUNTING TUBES.  
1. Begin by removing the headboard panel assembly (#2),  
if previously assembled to the bed, by pulling out the  
two Clevis Pins and two Hair Pin Clips that attach the  
main frame tubes to the headboard’s mounting tubes.  
Set all pins aside for later use.  
4. Insert the 1/4-20 x 2.50” Carriage Bolt (#5) into the vertical  
support as shown and fasten with a 1/4-20 Hex Head Lock  
Nut (#6). Both parts should be in your hardware bag.  
5. Assemble the headboard to the inside of the Trapeze  
Support so that the side of the headboard with the inserts  
is facing away from the Trapeze Support. From the outside,  
insert the four 1/4-20 x 1.75” long Machine Screws (#4)  
from your hardware bag through the holes in the Trapeze  
Support, screwing them into the inserts in the headboard.  
2. Detach the headboard panel from the mounting tubes  
by unscrewing the four 40mm Allen Head Bolts that  
hold the board to the tubes (these will be replaced  
with the longer Phillips Head Pan Screws from your  
trapeze adaptor hardware bag). Do not remove the  
four 1/4-20 inserts (#3) from the board. They will be  
used again to mount the board to the trapeze support.  
6. OPTIONAL - FOR TMI TRAPEZE BAR WITH SLEEVE:  
Insert the Sleeve into the vertical support with the slot on  
the bottom. Rotate the sleeve until it falls into place over  
the carriage bolt you inserted in Step 4. This will lock it into  
the proper position. (NOTE: The top of the Sleeve should  
sit flush with the top of the vertical support on the  
Trapeze Support Adaptor.)  
3. Assemble the Trapeze Support (#1) to the main frame  
tubes using the two Clevis Pins and Hair Pin Clips you  
removed in Step 1. The first holes in the Trapeze tubes  
must line up with the third hole from the end in the main  
frame tubes as shown.  
24  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
OPTIONAL BATTERY BACKUP REPLACEMENT CONFIGURATIONS  
Configuration 1:  
Replacing Single Battery  
Cable for a “Portable”  
Battery Backup Unit  
1. Replacing your existing battery backup  
cable (Part 999-0711-001) is fairly simple.  
First, if your external battery unit is hooked  
up, unplug it from the foot-end battery  
cable plug (#1).  
2. Carefully cut all cable ties (#2).  
3. Unplug the existing cable from the Control  
Box (#3). Plug in the new cable and run  
along same path as previous cable.  
4. Use cable ties to re-secure to tie rod and  
mounting brackets (#2).  
Configuration 2:  
Replacing Mounted Battery Backup Unit (includes attached battery cable)  
2. Using a 3/8” socket wrench, remove the two Nylon Lock Nuts  
(#6 - Part 100-6725-003) that secure the Battery Mounting  
Bracket (#5 - Part 999-0831-001) to the welded Main Frame  
Bracket (#4). Remove the bolts (#7 - Part 100-5225-009) and  
set aside.  
1. If your APS bed came with a mounted Battery  
Backup Pack (Part 999-0711-003) you must  
first unplug the power cord from the wall unit,  
then unplug the attached battery cable from  
the Control Box. Carefully cut all cable ties  
along the battery cable.  
3. Once the Bracket and Battery Pack (#9) are free, remove the  
pack from the bracket by unscrewing the four Phillips Head  
Screws (#8 - Part 100-5414-001). Place the new battery pack  
on the bracket and reverse the process to reattach bracket  
and cable.  
25  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
TROUBLESHOOTING - NOTHING WORKS - NO POWER  
1. NOTHING WORKS - NO POWER  
a. Check to make sure you have power coming from your  
outlet. Unplug the power cable from the outlet and test  
the outlet by plugging in a lamp or similar portable device.  
OUTLET WORKS = move to Step b.  
Test Outlet for Power.  
Unplug Power Cord  
and check entire cord  
to make sure it is not  
pinched, frayed, or  
damaged.  
b. Replug the power cord into the outlet -  
NOTHING WORKS - UNPLUG THE POWER CORD  
and check the power cord from the head end to the control  
box, making sure it is not pinched, frayed, or damaged  
in any way.  
- POWER CORD IS PINCHED = With cord unplugged,  
try to move the bed part slightly to release the pinched  
cord. If you can release the cord, re-plug the power  
cord into the wall outlet and test the bed.  
3-Prong Power Cable  
999-0775-208  
2-Prong Power Cable  
999-0775-206  
- BED WORKS = Make sure the cord is not frayed or  
exposed. If it is OK, you should not have to replace.  
- POWER CORD IS DAMAGED = Cut cable ties and  
immediately replace Power Cord.  
IF 3-PRONG (OR OPTIONAL 2 PRONG) POWER CORD  
IS DAMAGED, REPLACE IMMEDIATELY!  
ALWAYS UNPLUG THE POWER CORD BEFORE DOING  
ANY ELECTRICAL WORK ON YOUR BED.  
!
TROUBLESHOOTING STAFF CONTROL - BLINKING LIGHTS  
c. If you have a spare staff control assembly, plug the  
1. STAFF CONTROL LIGHTS ARE BLINKING  
cable into the T-Cable at the foot end and test.  
NOTHING ON STAFF CONTROL WORKS  
- WORKS AND LIGHTS NO LONGER BLINK =  
Replace the Staff Control Assembly in your foot-  
board. (See page 29 for dis-assembly. Please  
see page 32 for order number.)  
BUT PENDANT WORKS  
a. This could mean that your staff control is not getting  
enough power from the Control Box.  
b. Unplug the power cord from the wall outlet. Unplug  
the staff control cable from the T-Cable at the foot  
end of the bed.  
- DOESN’T WORK & LIGHTS STILL BLINKING =  
Replace the Control Box. (Please see Page 31  
for order number).  
26  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
NOTHING ON STAFF CONTROL WORKS BUT PENDANT WORKS  
a. Check the Staff Control connections at the foot end. Is the  
Staff Control Cable plugged securely into the T-Cable? Also  
check if the other end of the T-Cable is plugged securely  
into the Control Box (or Underbed Light). If the STAFF  
CONTROL STILL DOESN’T WORK = go to step b.  
b. Unplug the Power Cord from your wall outlet. Remove the  
Cable Cover (Shroud) on the inside of the footboard and  
locate the single terminal end that plugs directly into the  
back of the Staff Control Panel. Make sure it is properly  
seated into the panel. Plug in the power cord and test the  
staff control. If the STAFF CONTROL STILL DOESN’T  
WORK = Replace the Staff Control Assembly. (See page  
29 for dis-assembly. See page 32 for order information.)  
TROUBLESHOOTING PENDANT & DOUBLE T-CABLE  
If orange lock icon lit, press key symbol and function to unlock.  
A
1. STAFF CONTROL WORKS BUT PENDANT/  
HAND CONTROLLER FUNCTION DOESN’T  
a. Check to see if the functions are locked out on the Staff  
Control panel - if orange lock icons show on panel then  
Make sure plugs at foot  
end are fully seated  
that function is locked out. Press the Key and Function  
B
(i.e. Head) button at the same time until the orange lock  
icon goes out. Test the function. If the PENDANT STILL  
DOESN’T WORK = go to step b.  
and cap locked on.  
B
Make sure plugs  
are fully seated  
(O-ring does  
not show).  
b. Check the connections at the seat pan (Pendant/Hand  
Controller to T-Cable) and Staff Control Cable connection  
to the T-Cable at the foot end making sure plugs are fully  
engaged. Also check the T-Cable connection at the Control  
Box. If you have an Underbed Light feature, make sure all  
cables are secure. If the PENDANT STILL DOESN’T  
WORK = move to step c.  
Unplug Staff Control  
Cable from T-Cable  
at foot end.  
D
C
Unplug  
Power  
c. UNPLUG THE POWER CORD FROM THE WALL OUTLET.  
Plug  
before  
working  
on bed  
electonics.  
d. Unplug the Pendant/Hand Controller plug from the end of  
the T-Cable on either side of the beds Seat Pan. Unplug  
the Staff Control Cable from the end of the T-Cable at the  
foot end of the bed and unplug the other end from the  
Control Box.  
Plug Pendant Cable directly into HB port on Control Box  
(or into port 2 in Underbed Light Box) and test.  
e. Plug the Pendant/Hand Controller plug directly into the HB  
port on the Control Box, making sure it is fully seated.  
Plug in the Power Cord and test the Pendant.  
E
E
- ALL BUTTONS WORK = Replace the T-Cable  
(See page 32 for order information.)  
- NOTHING WORKS = Replace the Hand Controller.  
(See page 31 for order information.)  
27  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
TROUBLESHOOTING MOTOR CABLES & CONTROL BOX  
1. HEAD, FOOT, OR HI/LO FUNCTION  
NOT WORKING  
a. Check plugs to make sure they are firmly seated in all ports  
on your Control Box and are plugged into the correct ports  
(See diagram at right).  
POWER PLUG  
b. Check all cables to make sure they are not frayed, pinched,  
or damaged in any way. If any cable is damaged, UNPLUG  
YOUR BED’S POWER CORD FROM THE WALL OUTLET  
and REPLACE THE CABLE AND/OR THE ELECTRONIC  
COMPONENT.  
BATTERY  
c. HEAD DECK NOT RAISING - Unplug the power cord. Switch  
the head and foot plugs (ports 1 and 3). Re-plug the power  
cord and press the FOOT button to test.  
- HEAD WORKS = Control Box head port bad.  
Replace box (See page 31 for order information.)  
- DOESN’T WORK = Replace the Head Motor.  
(See page 31 for order information.)  
d. FOOT AND KNEE DECKS NOT RAISING - Unplug the  
power cord. Switch the foot and head plugs (ports 3 and 1).  
Re-plug the power cord and press the HEAD button to test.  
- FOOT WORKS = Control Box foot port bad.  
Replace box (See page 31 for order information.)  
2. BED LOOKS UNLEVEL - HILO MOTORS  
ARE NOT IN SYNC.  
- DOESN’T WORK = Replace the Foot Motor.  
(See page 31 for order information.)  
a. Press the Hi/Lo “DOWN” button repeatedly on either  
the Staff Control or Pendant/Hand Controller until  
the bed goes completely down and rests in its full  
horizontal position. The two HI/LO motor shafts  
should be completely retracted inside the motors.  
e. BED DOES NOT GO UP AND DOWN IN HORIZONTAL  
POSITION - If you press the Up or Down Hi/Lo button and  
you hear a click or buzz but nothing happens = Check your  
Hi/Lo Motors by touching them & pushing the Hi/Lo buttons.  
- YOU FEEL NOTHING = possible bad motor - go to the  
third step below.  
QUICK REFERENCE  
- HI/LO & FOOT WORKS, BUT HEAD DOES NOT:  
Switch Head & Foot Motor Cable at the Control Box  
and test (See Step C in left column).  
- YOU FEEL SLIGHT VIBRATION = Motor OK.  
- Unplug the Power Cord = Switch the two Hi/Lo plugs  
(ports 4 and 2) in the Control Box. Re-plug the power cord  
and press the same Hi/Lo button, then touch the same  
non-vibrating motor.  
- HI/LO & HEAD WORKS, BUT FOOT DOES NOT:  
Switch Foot & Head Motor Cable at the Control Box  
and test (See Step D in left column).  
i. MOTOR NOW VIBRATES = Bad port in the  
Control Box - Replace box. (See page 31 for  
order information.)  
- HEAD & FOOT WORKS, BUT HI/LO DOES NOT:  
a. Switch Head and Foot Hi/Lo Motor Cables at the  
Control Box and test (See Step E in left column).  
b. Test to see if Hi/Lo Motors need to be re-aligned  
(See Step F above).  
ii. MOTOR DOES NOT VIBRATE = Replace  
the bad Hi/lo Motor. (See page 31.)  
28  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
TROUBLESHOOTING/REPLACING YOUR APS STAFF CONTROL  
REMOVING AND REPLACING THE  
STAFF CONTROL ASSEMBLY  
6WDIIꢄ&RQWURO  
$VVHPEO\  
1. UNPLUG THE POWER CORD FROM  
THE WALL OUTLET.  
2. Unscrew the four Phillips Screws (#1) that attach the  
Shroud (cable cover) to the footboard and set aside.  
ꢀꢁꢂꢃꢄ  
6WDIIꢄ&RQWURO  
3. At the foot end of the bed, unscrew the retaining  
cap on the T-Cable (#3) and unplug the two cables.  
%HYHO  
)RRWꢄ%RDUG  
ZLWKꢄFXWRXW  
6FUHZV  
4. Unscrew the two remaining Phillips Screws (#2)  
from the inside of the footboard that attach the  
Staff Control Assembly bevel to the board. Set  
aside screws for later re-assembly.  
ꢀꢁꢂꢃꢄ  
6WDIIꢄ&RQWURO  
%HYHOꢄ6FUHZV  
5. Remove the OLD Staff Control Assembly and fish  
the old cable out through the large hole.  
ꢀꢁꢅꢃ  
6KURXGꢄ6FUHZV  
6. Feed the cable of the NEW Staff Control  
Assembly from the outside through the  
large hole and insert the bevel into the  
cutout on the outside of the footboard.  
ꢀꢁꢆꢃꢄ  
ꢀꢁꢇꢃꢄ  
ꢀꢁꢈꢃꢄ  
7. Secure the Staff Control by  
re-inserting the two Phillips  
Screws (#2) you removed  
in Step 4.  
,03257$17ꢉꢄ5XQꢄDꢄ&DEOHꢄ  
7LHꢄWKURXJKꢄWKHꢄHQGꢄKROHꢄLQ  
WKHꢄPRXQWLQJꢄWXEHꢊꢄORRSꢄWKH  
H[FHVVꢄFDEOHꢊꢄUXQꢄWKHꢄWLH  
DURXQGꢄFDEOHꢊꢄDQGꢄSXOOꢄWLJKW  
WRꢄVHFXUHꢋꢄ&XWꢄRIIꢄH[FHVV  
FDEOHꢄWLHꢋ  
8. Re-position the Shroud on  
the inside of the footboard  
and secure with the four  
Phillips Screws (#1) you  
removed in Step 2.  
9. Plug the T-Cable (#5) into the Staff Control  
Cable (#4), making sure the phone jack plug  
is seated correctly into the Staff Control cable plug.  
Screw on the retaining cap (#3) to secure.  
10. Run a Cable Tie through the large end hole on  
the right-hand footboard mounting tube.  
)HPDOHꢄHQG  
RIꢄ6WDIIꢄ&RQWURO  
&DEOH  
11. Loop the excess Staff Control Cable. Run the  
cable tie around the cable and pull tight to secure.  
Cut off any excess Cable Tie.  
(QGꢄRIꢄ'RXEOHꢄ7ꢌ&DEOH  
ꢄDWꢄIRRWꢄHQGꢄꢌꢄ*UH\  
ꢀꢍ7ꢍꢄUXQVꢄRQꢄHLWKHU  
VLGHꢄRIꢄ6HDWꢄ3DQ  
IRUꢄDWWDFKLQJꢄ\RXU  
+DQGꢄ&RQWUROOHUꢋꢃ  
/RFNLQJ  
(QGꢄ&DS  
29  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
RECOMMENDED MAINTENANCE AND INSPECTION  
} Maintenance Inspection of All  
Components (Receipt of shipment)  
} Electrical Inspection of Control Box  
(6 months)  
• Check the external power cord that plugs into the control  
box for any chafing cuts or wear.  
• Make sure all parts/components are included  
(Please see page 9 - “Unpacking Your Bed”).  
• Make sure all attaching hardware is securely tightened.  
• Check all electrical connections for any wear or fractures.  
• Check all bed components for obvious damage.  
• Inspect the Power Supply Cord for any cuts and/or damage.  
• Check your external or mounted battery (see page 24)  
Replace if needed.  
• Check to see all actuator cords are routed and connected  
properly to the control box.  
} Electrical Inspection of Pendant  
and Staff Control (6 months)  
} Mechanical Inspection of Assemblies  
(6 months)  
• Inspect all welds on the sleeping surface, frame, and base  
assemblies for stress fractures.  
• Check the pendant cord for any chafing, cuts, or wear.  
• Check all pendant functions . . .  
• Inspect all fasteners for wear and looseness.  
P
P
P
P
P
Head raises and lowers properly  
Foot raises and lowers properly  
• IMPORTANT: Lubricate all pivot points, actuator/motor  
clevis pins, and control arm clevis pins as needed. White  
Lithium Grease is recommended.  
Entire bed raises and lowers properly  
Bed moves to Chair position properly  
Bed moves to Trendelenberg and  
} Mechanical Inspection of Casters  
& Pedal Locking Mechanism  
(3 months)  
Reverse Trendelenberg positions properly  
• Check to make sure each button and associated function  
work properly (i.e. the head section raises when the Head  
Up button is activated).  
• Check the pedal locking mechanism to make sure it engages  
and disengages properly.  
} Electrical Inspection of Actuators/Motors  
(6 months)  
• Check the bottom foot pads on the foot end casters, making  
sure they are clean of debris and undamaged. Replace if needed.  
• Check all casters to ensure that they roll properly and  
are unobstructed.  
• Check the actuator/motor cords for any chafing, cuts, or wear.  
• Check the range of movement on all motors to ensure they  
do not bind in the Full Up and Full Down positions.  
WARRANTY  
- 15 Years on Frame  
- 5 Years on Motors and Control Box  
(longest motor warranty in the industry)  
- 3 Years on Other Electronics  
The warranties contained herein contain all the representation and warranties with respect to the subject matter  
of this document, and supersede all prior negotiations, agreements and understanding with respect thereto.  
The recipient of this document hereby acknowledges and represents that it has not relied on any representation,  
assertion, guarantee, warranty, collateral contract or other assurance, except those set out in this document.  
30  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
SERVICE/REPLACEMENT PART RECORD (Page 1)  
DECKING  
Part Number  
999-0844-920SP  
999-0844-930SP  
999-0844-940SP  
Description  
QTY  
1 Deck  
1 Deck  
1 Deck  
1 Pkg  
Order Date  
Price  
Head/Back Deck & Hardware Pack  
OR - Foot Deck & Hardware Pack  
OR - Knee Deck & Hardware Pack  
Deck Hardware Pack - INCLUDES . . .  
Clevis Pins  
4
5
2
1
5/16” Retaining Rings (Deck Brackets)  
12mm Retaining Ring (Cntrl Arms to Decks)  
Installation Instruction Sheet  
MOTORS (ACTUATORS)  
Description  
Part Number  
999-0831-051SP  
999-0822-052SP  
999-0822-053SP  
QTY  
1 Motor  
1 Motor  
1 Motor  
1 Pkg  
Order Date  
Price  
Hi/Lo Motor (Cable not included) & Hardware Pack  
OR - Head Motor (Cable not included) & Hardware Pack  
OR - Foot Motor (Cable not included) & Hardware Pack  
Motor Hardware Pack - INCLUDES . . .  
2
2
2
8
1
12mm Retaining Rings  
Hair Pin Clips  
Snap Rings  
Standard Cable Ties  
Installation Instruction Sheet  
999-0806-200  
999-0806-202  
Head and Hi/Lo Motor Cables  
Foot Motor Cable  
1 per Motor  
1 Cable  
CONTROL BOX & POWER CABLES  
Part Number  
999-0831-300  
999-0775-208  
999-0775-206  
Description  
QTY  
1 Box  
Order Date  
Order Date  
Price  
Price  
Linak Control Box and Clip  
Standard 3 Prong Power Cable Ground Wire  
Optional 2 Prong Power Cable  
1 Cable  
1 Cable  
PENDANT/HAND CONTROLLER  
Part Number  
999-0806-301  
999-0791-000  
999-0806-305  
Description  
QTY  
Backlit Pendant with Underbed Light button  
Pendant Holster  
1 Pendant  
1 Holster  
1 Pendant  
Non-Backlit Pendant (no Underbed Light button)  
31  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
SERVICE/REPLACEMENT PART RECORD (Page 2)  
DOUBLE T-CABLE  
Part Number  
Description  
QTY  
1 Cable  
1 Pkg  
Order Date  
Order Date  
Order Date  
Price  
Price  
Price  
999-0806-200SP  
Double “T” Control Cable  
Deck Hardware Pack - INCLUDES . . .  
Split Stem Cable Retainer  
Push Cable Ties  
2
4
8
1
Standard Cable Ties  
Installation Instruction Sheet  
UNDERBED LIGHT  
Description  
Part Number  
QTY  
1 Motor  
1 Pkg  
999-0806-220SP  
Underbed Light Unit  
Motor Hardware Pack - INCLUDES . . .  
Adaptor Cable - 999-0806-212  
Standard Cable Ties  
12  
4
1
Attachment Material  
1
Installation Instruction Sheet  
STAFF CONTROL  
Description  
Part Number  
999-0831-901  
ZL831000  
QTY  
1 Unit  
1 Pkg  
Switch Pad with Bezel & Cables  
Staff Control Service Pack - INCLUDES . . .  
1
1
6
3
1
Switch Pad with Bezel & Cables  
Shroud (Cable Cover)  
#6 Phillips Head Truss Screws  
4” Nylon Cable Tie  
PLEASE SEE  
PAGE 11  
FOR  
INSTALLATION  
INSTRUCTIONS  
Installation Instruction Sheet  
CAPS & PLUGS  
Description  
Part Number  
QTY  
Order Date  
Price  
NOTE: CAPS & PLUGS CAN ONLY BE PURCHASED IN SETS OF 12  
100-4200-004PK  
100-4700-017PK  
100-4700-018PK  
100-4715-011PK  
999-0775-001PK  
100-4762-002PK  
100-4738-005PK  
12/pkg  
12/pkg  
12/pkg  
12/pkg  
12/pkg  
12/pkg  
12/pkg  
1.25” x 1.25” Square Caps Pack  
1.00” x 2.00” Rectangular Caps Pack  
1.25” x 2.00” Rectangular Caps Pack  
1.50” Round End Caps Pack (cover arm bearings)  
Half Moon End Caps Pack (caster bases)  
5/8” Round Plugs Pack (both sides of foot deck)  
3/8” Round Plugs (on main frame rails - foot end)  
FOR GENERAL CAP/PLUG DIAGRAMS PLEASE SEE PAGES 36 & 37  
32  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
SERVICE/REPLACEMENT PART DIAGRAM - ELECTRONICS  
FOOT-END  
HI/LO MOTOR  
END OF  
BATTERY CABLE  
(BLACK)  
CONTROL BOX -  
ATTACHED TO MOTOR  
FOOT-END  
OF T-CABLE  
(GREY)  
HEAD-END  
HI/LO MOTOR  
LONGER SIDE OF T-CABLE  
WITH PUSH CABLE TIE IN  
SMALL HOLE  
FOOT  
END  
FOOT/KNEE  
MOTOR  
LONG “T” OF T-CABLE  
AND BATTERY CABLE  
SECURED TO ROD  
AND MAIN FRAME  
BRACKET WITH  
CABLE TIES  
FOOT MOTOR  
SHORTER SIDE OF T-CABLE WITH  
PUSH CABLE TIE IN SMALL HOLE  
HEAD  
END  
CABLE PROTECTION GROMMET  
IN SEAT PAN KEY HOLE  
POWER CABLE  
PLUG AT HEAD END  
POWER CABLE SECURED TO ROD  
WITH CABLE TIES  
1. TO ORDER A REPLACEMENT  
HI/LO MOTOR USE SERVICE PACK  
999-0831-051SP (Includes motor,  
mounting hardware & installation  
instruction sheet).  
2. TO ORDER A REPLACEMENT  
3. TO ORDER A REPLACEMENT  
FOOT MOTOR USE SERVICE PACK  
999-0822-053SP (Includes motor,  
mounting hardware & installation  
instruction sheet).  
HEAD MOTOR USE SERVICE PACK  
999-0822-052SP (Includes motor,  
mounting hardware & installation  
instruction sheet).  
4. TO ORDER A REPLACEMENT  
CONTROL BOX PLEASE USE  
999-0831-300 (Includes Control  
Box and mounting clip. Power  
Cord must be ordered separately  
- see page 31 for order numbers)  
5. TO ORDER A REPLACEMENT  
DOUBLE T-CABLE USE SERVICE  
PACK 999-0806-200SP (Includes  
T-Cable, cable ties, grommets &  
installation instruction sheet).  
6. TO ORDER A REPLACEMENT  
BATTERY MOTOR USE SERVICE  
PACK 999-0711-001SP (Includes  
battery cable, long cable ties, and  
installation instruction sheet).  
WHEN ORDERING REPLACEMENT PARTS WITH CUSTOMER SERVICE,  
PLEASE HAVE YOUR BED’S SERIAL NUMBER AVAILABLE  
TO CONFIRM WHETHER THE PART IS COVERED UNDER WARRANTY.  
(SEE PAGE 5 FOR LOCATION OF SERIAL NUMBER ID LABEL.)  
33  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
SERVICE/REPLACEMENT PART DIAGRAM - DECKING  
HEAD DECK  
TO MAIN FRAME  
SEAT PAN BRACKETS  
DETAIL OF DECK  
ATTACHMENT  
(Similar for all decks)  
KNEE DECK  
TO MAIN FRAME  
SEAT PAN BRACKETS  
FOOT DECK  
TO KNEE DECK  
MOUNTING BRACKETS  
DETAIL OF EDEMA  
RATCHET RE-ATTACHMENT  
(Use existing ratchets  
and hardware)  
1. TO ORDER A REPLACEMENT  
2. TO ORDER A REPLACEMENT  
KNEE DECK USE SERVICE PACK  
999-0844-940SP (includes deck,  
mounting hardware & installation  
instruction sheet).  
3. TO ORDER A REPLACEMENT  
FOOT DECK USE SERVICE PACK  
999-0844-930SP (includes deck,  
mounting hardware & installation  
instruction sheet). Re-use existing  
edema ratchets & ratchet hardware.  
HEAD DECK USE SERVICE PACK  
999-0844-920SP (includes deck,  
mounting hardware & installation  
instruction sheet).  
WHEN ORDERING REPLACEMENT PARTS WITH CUSTOMER SERVICE,  
PLEASE HAVE YOUR BED’S SERIAL NUMBER AVAILABLE  
TO CONFIRM WHETHER THE PART IS COVERED UNDER WARRANTY.  
(SEE PAGE 5 FOR LOCATION OF SERIAL NUMBER ID LABEL.)  
34  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
SERVICE/REPLACEMENT PART DIAGRAM -  
MAIN FRAME CAPPING  
HEAD  
END  
FOOT  
END  
CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY  
(PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOSE  
ANY CAPS/PLUGS ON YOUR BED - SEE BOTTOM OF PAGE 32 FOR ORDER NUMBERS)  
35  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
SERVICE/REPLACEMENT PART DIAGRAM -  
CAPPING ON HEAD, KNEE, AND FOOT DECKS  
CAPPING ON ARMS & LEGS  
2X 100-4700-018  
End Caps  
2X 100-4700-018  
Rectangular End Caps  
4X 100-4715-011  
Round End Caps  
2X 100-4700-021  
Rectangular End Caps  
CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY  
(PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOSE  
ANY CAPS/PLUGS ON YOUR BED - SEE BOTTOM OF PAGE 32 FOR ORDER NUMBERS)  
36  
999-0831-190A Service Manual  
Basic American Medical Products, a division of GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Zenithaps seRies  
NOTES:  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
_________________________________________________________________________________  
37  
Basic American Medical Products, a division of GF Health Products, Inc.  
999-0831-190A Service Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
336 Trowbridge Dr. • Fond du Lac, WI 54937  
Customer Service: 1.800.365.2338 • Fax: 920.929.8213  
© 2012 GF Health Products, Inc.  
All Rights Reserved  
Basic American Medical Products and Zenith  
are registered trademarks of GF Health Products, Inc.  
GF121699RevA12  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Gravely Lawn Mower 31217 User Manual
Haier Washer A6 707 User Manual
Harbor Freight Tools Nail Gun 93760 User Manual
Hearth and Home Technologies Outdoor Fireplace GO24MTCH User Manual
Honeywell Home Security System 4110XM User Manual
HP Hewlett Packard Calculator 35s User Manual
HP Hewlett Packard Printer 1320 Series User Manual
HP Hewlett Packard Satellite Radio 8743A User Manual
Hustler Turf Lawn Mower Lawn Mowe User Manual
Image Treadmill IMTL119050 User Manual