LF1031FP HYDRAULIC PATIENT LIFT
WITH FOOT PEDAL
USER MANUAL
LF1031FP-INS-LAB-RevA11
© 2011 GF Health Products, Inc.
Read this manual before operating your Patient Lift.
Save this manual for future use.
The most current version of this manual can be found online at
Download from Www.Somanuals.com. All Manuals Search And Download.
1 INTRODUCTION
Congratulations on your purchase of a Lumex patient lift. The following pages will provide you
with important safety and operating instructions on the use of patient lifts, slings, and accessories,
as well as maintenance and warranty information. Read this manual carefully before operating
your patient lift and refer to it as often as needed. Consult your Lumex dealer and/or healthcare
professional with any questions or concerns regarding safe and effective techniques for operating
your patient lift.
The Lumex Patient Lift is intended to aid in the transfer of an individual who is unable to indepen-
dently transfer between bed, chair, wheelchair, and/or commode.
DO NOT attempt any transfer without prior approval of the patient’s healthcare provider. With
their assistance, select the sling that best serves the needs, abilities, comfort and safety of the per-
son being lifted. It is important that you use correct slings adapted to the body shape and weight
and ability of the patient and the specific type of transfer needed.
Upon receipt of the patient lift, and prior to each use, inspect it closely to ensure that nothing is
damaged, there are no loose or broken parts, that all patient lift parts are in the proper place, etc.
Immediately replace any worn or broken components. Your patient lift should receive maintenance
on a regular schedule and should be inspected daily for proper operation. Contact your Lumex
dealer for service and repairs.
Before lifting anyone, thoroughly read the instructions in this manual. Ideally, observe a team of
trained experts performing the lifting procedures. The operator should then practice the entire
lifting procedure several times with proper supervision and with a capable individual acting as
patient. It is extremely important to know how a patient will feel in the patient lift. We suggest that
the operator try all the various lifting slings and equipment until the operator is confident with
how the patient lift operates. The lifting procedures should be explained to the patient before the
actual lifting. Remember that safety requires the constant attention of the lift occupant and the
operator. Use extreme care and common sense when operating a patient lift.
This guide covers the use of the LF1031FP Hydraulic Patient Lift. Keep in mind that the basic safety
procedures included in this manual are to be used as a guide only. You may find it necessary to de-
velop your own unique methods for safely utilizing the patient lift. Again, consult your healthcare
professionals for their recommendations and never hesitate to ask for their assistance.
Info: Before attempting to assemble the patient lift, refer to the section titled “ASSEMBLY IN-
STRUCTIONS” contained elsewhere in this manual. Contact your Lumex dealer if additional
assistance is needed.
Info: In this manual, the person being lifted is referred to as the “patient” or “occupant”
and the person operating the patient lift is called the “operator” or “attendant”.
Please note the following special statements, used throughout this manual, and their significance:
WARNING: Indicates a potential hazard situation or unsafe practice that, if not avoided,
could result in death or serious personal injury.
CAUTION: Indicates a potential hazard situation or unsafe practice that, if not avoided,
could result in minor or moderate personal injury.
s NOTICE: Indicates a potential hazard situation or unsafe practice that, if not avoided,
could result in product or property damage.
Info: Provides application recommendations or other useful information to ensure that you
get the most from your product.
1031FP-INS-LAB-RevA11
3
Download from Www.Somanuals.com. All Manuals Search And Download.
2 SAFETY PRECAUTIONS
IMPORTANT: Before using patient lift, please read and adhere to the following safety precautions
and warnings. Failure to do so could result in serious personal injury or damage to your patient
lift.
Always consult your healthcare professional to determine safe methods most suitable for your
individual abilities. Protect yourself, your attendant and patient lift by having it serviced regularly.
If you experience any malfunction, contact your Lumex dealer immediately, as a hazardous condi-
tion could result, causing personal injury or damage to your patient lift.
Periodic inspection, adjustment and replacement of worn parts are necessary to provide years of
excellent service. Refer to CARE AND MAINTENANCE section of this manual.
Maintenance MUST be performed by qualified personnel ONLY.
WARNINGS
WARNING: Important! Read and understand these instructions before assembling or
using the LF1031FP. If you do not understand any part of these warnings, cautions or
instructions, contact a healthcare professional for direction in the use of this product.
If the Lumex Patient Lift is not properly assembled, personal injury and damage to the
Lumex Patient Lift could result.
WARNING: If components are damaged or missing, contact your dealer immediately. DO
NOT use substitute parts. Use only Lumex replacement parts. The use of non-Lumex re-
placement parts could cause personal injury, property damage, and void the warranty.
WARNING: Unauthorized modification of your patient lift could cause personal injury,
property damage, and void the warranty.
WARNING: The LF1031FP Hydraulic Patient Lift maximum weight capacity is 400
pounds. DO NOT exceed the maximum weight capacity.
WARNING: GF Health Products, Inc. assumes no responsibility for any damage or injury
caused by improper assembly or use of this product.
WARNING: Check all parts for shipping damage before using. In case of damage, DO
NOT USE the equipment. Contact the carrier or your authorized Lumex dealer for further
instructions.
WARNING: Patient lift may tip over if used incorrectly. Read and adhere to the operating
instructions prior to lifting anyone.
WARNING: GF Health Products, Inc. strongly recommends that two caregivers take part
in the lifting process.
WARNING: When transferring a patient, always keep the patient centered over the base.
WARNING: This patient lift is designed for transfer only. It is not intended to be used for
transporting or moving a patient from one location to another.
WARNING: Ensure that lifting sling loops are correctly attached to the hooks to prevent
the patient from sliding or falling out of the sling, which could result in personal injury.
Never use the patient lift to lift or transfer anything other than a person.
WARNING: Do not lock the brakes or block the wheels when lifting. The casters must be
FREE to roll to allow the patient lift to stabilize itself when the patient is initially lifted
from a chair, bed, or any stationary object.
WARNING: To ensure stability while transferring a patient, the patient lift legs should be
locked in the maximum open position.
1031FP-INS-LAB-RevA11
4
Download from Www.Somanuals.com. All Manuals Search And Download.
WARNING: Replace any worn or broken parts immediately! Torn, cut, frayed, or broken
slings can fail, resulting in personal injury. Use only slings that are in good condition.
WARNING: The hydraulic pump is sealed at the factory. If service is required, the unit
must be returned to the factory for repair. DO NOT attempt to open the hydraulic pump
or obtain local service; this will void the warranty and may result in damage to the unit.
Consult your Lumex Dealer for further information.
WARNING: Use extreme care, caution, and common sense when utilizing a patient lift.
Special attention must be given to patients who are disoriented and cannot cooperate
while being lifted, such as comatose, agitated, spastic, or mentally confused persons.
Consult their healthcare professional for assistance in lifting these types of patients.
WARNING: Lumex specifically disclaims responsibility for any bodily injury or property
damage which may occur during any use which does not comply with federal, state or
local laws or ordinances.
1031FP-INS-LAB-RevA11
5
Download from Www.Somanuals.com. All Manuals Search And Download.
3 FEATURES
boom (height-adjustable)
mast
adjustment handle
(boom height adjustment)
6-point spreader bar
(suspension)
hydraulic pump
push handle
pump knob
(pump locking and unlocking)
foot pedal
(base width adjustment)
locking
casters
base (width-adjustable)
swivel
casters
LF1031FP Hydraulic Patient Lift with Foot Pedal
1031FP-INS-LAB-RevA11
6
Download from Www.Somanuals.com. All Manuals Search And Download.
4 HANDLING PROCEDURES
SHIPPING AND TRANSPORTATION INSTRUCTIONS
If the patient lift is to be reshipped by common carrier, it should be packed in the same carton in
which it was received.
Unpacking
1. Check for any obvious damage to the carton or its contents. If damage is evident, notify the car-
rier or your authorized Lumex Dealer.
2. Remove all loose packing from the carton.
3. Carefully remove all the components from the carton.
Info: Unless the patient lift is to be used immediately, retain boxes, containers and packing
material for use in storing until use of patient lift is required.
Inspection
1. Examine the exterior of the patient lift for nicks, dents, scratches or other damage.
2. Inspect all components.
Storage
1. Store the repackaged patient lift in a dry area.
2. DO NOT place other objects on top of the repackaged patient lift.
1031FP-INS-LAB-RevA11
7
Download from Www.Somanuals.com. All Manuals Search And Download.
5 ASSEMBLY INSTRUCTIONS
Components are shown in illustration below and called out in list below illustration.
COMPONENTS
1
7
4
9
6
10
3
5
8
2
Components:
LF1031FP Hydraulic Patient Lift
with Foot Pedal
Item Qty Description
1
1
1
2
2
1
1
1
1
1
2
Mast/Boom/Pump assembly
2
Base
3
L-insert
4
Push handle
5
Spreader bar to boom attachment bolt and hardware
Spreader bar to boom attachment rod and cap
Spreader bar
6
7
8
Assembly wrench set
9
Mast locking pin
10
Mast attachment bolt
1031FP-INS-LAB-RevA11
8
Download from Www.Somanuals.com. All Manuals Search And Download.
ASSEMBLY
1. Remove the locking pin from the mast sleeve
located on the rear of the base.
remove locking pin
2. Install the L-inserts in the base.
install L-inserts
3. With the boom pointed in the same direction as
the base legs, insert the mast into the sleeve.
4. Insert the locking pin through the holes of the
sleeve and mast, and attach the locking cable to
the end of the pin.
install mast and locking pin
5. Screw the two mast attachment bolts into the
holes in the sleeve of the base.
install mast attachment bolts
1031FP-INS-LAB-RevA11
9
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Screw the push handles into the mast and tighten the
nuts.
install push handles
7. Insert spreader bar attachment rod into spreader bar.
Insert spreader bar rod into the U-bracket at the end of
the boom. Align holes. Insert spreader bar attachment
bolt and tighten.
install spreader bar
8. For shipping purposes, the boom is in its lowest posi-
tion. To elevate, turn the pump knob clockwise until
hand tight and operate the pump handle.
9. Operate the foot pedal to
adjust the base width.
elevate boom
The picture at near right
shows the base end op-
posite the foot pedal at
minimum width; the left
food pedal has been com-
pletely depressed.
The picture at far right
shows the base end op-
base at minimum width
base at maximum width
posite the foot pedal at
maximum width; the right
foot pedal has been com-
pletely depressed.
1031FP-INS-LAB-RevA11
10
Download from Www.Somanuals.com. All Manuals Search And Download.
6 OPERATING INSTRUCTIONS
LIFTING THE PATIENT
WARNING: Do not attempt to transfer a patient without prior approval of the patient’s
healthcare professional and without having read the instructions and practiced using
the patient lift.
WARNING: Use special care for those patients who cannot cooperate while being lifted.
WARNING: Ensure all bolts, hooks, and loops are securely fastened.
WARNING: Use only Lumex slings with Lumex patient lifts.
WARNING: DO NOT use torn, cut, frayed or broken slings as injury could result.
WARNING: Ensure the floor area is clear of any obstacle that could cause a caster to
stop and possibly tip the patient lift.
Info: Instructional videos that review the application of several sling styles and lifting sce-
Lifting the patient from a bed
1. With patient face-up in center of bed, roll patient onto their side away from attendant.
2. Place sling, folded lengthwise, halfway under patient with lower edge just beneath knees.
3. Roll patient onto their side toward attendant, pull sling to smooth it, and center patient on sling.
4. Pull the leg loops forward and under the thighs.
5. Cross the loops, one through the other.
6. Roll the patient lift under the bed, locating the spreader bar over the patient. Be careful not to
lower the spreader bar onto the patient. Center the spreader bar over the patient before lifting.
WARNING: Do not lock or block the patient lift casters when lifting. The casters must be
free to roll so that the patient lift can stabilize as the patient is lifted from the bed.
WARNING: To ensure stability while lifting or transferring a patient, the patient lift legs
must be locked in the maximum open position.
7. Hook the sling loops onto the spreader bar.
8. If the patient is in a hospital bed, raise the bed's head section to elevate the patient.
9. Raise the patient carefully. If possible, lower the bed. Reposition the patient as needed, using
the handle on the back of the sling, to ensure a safe and comfortable sitting position as the
patient is raised.
10. Raise the patient until the buttocks are just above the mattress. Lift the patient’s legs and turn
the patient to face the lifting column. Move the patient lift away from the bed. Ensure that the
patient's arms and legs clear the patient lift, bed, or any other obstacles.
Lowering the patient onto a bed
1. Raise the patient lift until the patient’s buttocks are above the surface of the bed. Ensure that
there are no obstacles underneath or around the bed. Move the patient lift under the bed. Ensure
the patient does not hit any portion of the bed and that the patient’s legs have cleared the bed.
2. Center the patient over the bed, and rotate the patient to face the foot section of the bed. SLOW-
LY release the pressure knob by turning counter-clockwise for gradual lowering of the patient.
Carefully lower the patient onto the bed. Support the patient’s head, if necessary, while being
lowered. Remove the sling by reversing the procedures used when lifting patient from bed.
1031FP-INS-LAB-RevA11
11
Download from Www.Somanuals.com. All Manuals Search And Download.
Lifting the patient from the floor
1. Position the patient lift behind the patient, placing a pillow over the base to protect and sup-
port the patient’s head and neck. Place the patient's head carefully on the pillow. The patient
lift should be positioned with the spreader bar centered over the patient.
2. Lower the spreader bar. Attach the sling to the spreader bar hooks.
3. Lift the patient up above the floor. Position the patient in the sling by pulling the handle on the
back of the sling.
4. Turn the patient to face the lifting column. Remove pillow from base and place the patient’s
feet on the base, straddling the column for extra stability.
Lowering the patient to the floor
1. Place a pillow on the base to protect the patient’s head and neck (this is unnecessary when us-
ing a sling with a head support).
2. Reverse the above procedures to lower.
Transferring the patient from a chair to a wheelchair or commode chair
WARNING: GF Health Products, Inc. recommends that two attendants be used when
transferring a patient to and from a wheelchair.
1. Fold the sling.
2. While maintaining support, gently lean the patient forward. Place the folded sling behind the
patient’s back and push it down until it touches the seat of the chair.
3. Use the loops to pull the leg sections of the sling forward and beneath the patient’s thighs.
Then cross the loops, one through the other.
4. Move the patient lift around to the front of the chair.
5. Hook all sling loops onto the spreader bar.
6. Carefully raise the patient.
7. The second attendant should now move the wheelchair into position.
8. Engage the wheelchair wheel locks to prevent movement.
WARNING: Before transfer, ensure wheelchair wheel locks are in locked position. Wheel-
chair wheel locks must be in a LOCKED POSITION before lowering the patient into the
wheelchair, or unexpected wheelchair movement could result, which may result in seri-
ous injury to both the patient and the attendants.
9. Position the patient over the wheelchair with their back against the back of the wheelchair.
10. SLOWLY lower the patient into the wheelchair.
11. With one attendant behind the wheelchair and the other operating the lift, the attendant be-
hind the chair should pull back on the handle or sides of the sling to place the patient into the
back of the chair. This will maintain a good center of balance and prevent the chair from tip-
1031FP-INS-LAB-RevA11
12
Download from Www.Somanuals.com. All Manuals Search And Download.
ping forward.
Removing the sling
Gently pull the leg sections to the side, out from under the patient’s thighs. Standing by the pa-
tient’s side, lean the patient forward while supporting the patient with one hand. Pull the sling up
from behind the patient’s back and position the patient in the chair.
WARNING: Pulling sharply on the sling may cause the patient to fall forward, resulting
in personal injury. Always support the patient by keeping one arm around the patient’s
shoulders when placing or removing the sling.
Info: If possible, the patient should be encouraged to take an active part in removing the
sling. The patient can lean forward, lift their thighs, and help in removing the leg sections.
Transferring the patient from a bed to a wheelchair or commode chair
1. For smooth, easy lifting, have the wheelchair or commode chair in ready position with wheel
locks in locked position.
2. With patient face-up in center of bed, roll patient onto their side away from attendant.
3. Place sling, folded lengthwise, halfway under patient with lower edge just beneath knees.
4. Roll patient to side toward attendant, pull sling to smooth it, and center patient on sling.
5. Pull the leg loops forward and under the thighs.
6. Cross the loops, one through the other.
7. With base of lift under bed, SLOWLY release pressure knob by turning counter-clockwise and
pull down boom. Attach sling to spreader bar.
8. Turn pressure knob clockwise until tight, and lift the patient by operating the pump handle.
9. Lift patient, maintaining patient's position facing attendant, until patient's feet swing easily off
the bed.
10. Position patient above wheelchair or commode chair. Lock lift brakes.
WARNING: Before transfer, ensure wheelchair or commode chair wheel locks are in
locked position. Wheelchair wheel locks must be in a LOCKED POSITION before lowering
the patient into the wheelchair, or unexpected wheelchair movement could result, which
may result in serious injury to both the patient and the attendants.
11. SLOWLY release pressure knob by turning counter-clockwise to gradually lower patient. During
descent, assist patient in attaining correct sitting posture.
12. When patient is seated, push down on boom to slacken sling loops. Patient can remain seated
on sling.
13. For commode chair use, adjust clothing before moving lift to straddle commode chair. Keep
sling loops taut and ensure that patient is in a comfortable position.
1031FP-INS-LAB-RevA11
13
Download from Www.Somanuals.com. All Manuals Search And Download.
7 CARE AND MAINTENANCE
Proper care and maintenance are essential to keeping your Lumex patient lift in a safe operating
condition. In addition to inspecting the unit before each use, periodic maintenance checks should
be done. We further recommend that this unit be serviced at least once annually by an authorized
dealer.
WARNING: It is extremely important that the patient lift be inspected before each use.
Ensure that all hardware and accessories are secure and that the actuator is function-
ing properly. Failure to do so could result in patient / attendant injury or damage to your
patient lift. Torn, cut, frayed or broken slings can fail, resulting in serious injury. Only
use slings in good condition. Inspect before each use. Destroy and discard old worn and
unusable slings.
When you believe a component or part is not functioning properly, immediately contact your au-
thorized Lumex dealer, as a potentially hazardous condition could exist.
BEFORE EACH USE
Before each use of the patient lift, ensure that:
• All nuts and bolts are tight
• The lift moves freely (with caster brakes off)
• Rear caster brakes operate properly
• Base adjustment mechanism operates easily
• Spreader bar clips are secured tightly
• The hydraulic pump operates properly
LUBRICATION
Your Lumex patient lift was designed with minimum maintenance in mind; however, a six (6)
month check and lubrication should ensure that your patient lift will remain in peak operating con-
dition. The casters MUST roll and swivel freely. Lightweight waterproof grease (such as automobile
lubricant) should be applied to the ball bearings of the casters every six (6) months. Apply grease
more frequently if the casters are exposed to periods of excess moisture. Lubricate all pivot points
with lightweight oil. Clean patient lift of all excess oil and grease.
Before use of the patient lift, and every two months thereafter, place a drop of oil on the following
areas:
• The point where mast and boom connect
• Spreader bar hooks
• Pump handle hinge
• Caster axles
• The point where pump and boom connect
1031FP-INS-LAB-RevA11
14
Download from Www.Somanuals.com. All Manuals Search And Download.
CLEANING
Regular cleaning is recommended. A soft cloth, dampened with water and a mild NON-ABRASIVE
detergent (household cleaner, soapy water, or hospital grade disinfectant) is all that is needed to
clean your patient lift. Automobile wax or furniture polish will help maintain the finish over a long
period of time.
s NOTICE: DO NOT wash patient lift under water pressure or steam clean.
CHECKING FOR DAMAGE AND WEAR
It is extremely important to frequently check the areas of the patient lift that are under a great deal
of stress for damage and wear. These areas include slings, spreader bars, spreader bar suspension
and mounting hardware, pump mounting hardware and all pivot points. Inspect for signs of crack-
ing, fraying, deformation, and any loose components
s NOTICE: Replace any defective parts immediately, and make certain that the patient lift
is not used until repairs are made. Contact your authorized Lumex dealer for replace-
ment parts.
CARE OF SLINGS
Lumex slings are washable at 150° F and autoclavable at 185° F for 30 minutes. The head support's
plastic insert should be removed before washing. The slings can be dried by using a dryer on LOW
heat or by hanging.
s NOTICE: DO NOT bleach slings!
1031FP-INS-LAB-RevA11
15
Download from Www.Somanuals.com. All Manuals Search And Download.
INSTALLATION OF 1010S005 HYDRAULIC PUMP
Removal of Existing Pump
1. Follow the instructions in this
upper boom
bracket
lower mast
bracket
boom
section to remove the existing
pump, shown at near right. If
installing pump for the first time,
proceed to INSTALLATION OF
NEW PUMP.
existing pump
mast
2. Remove the top of the exist-
ing pump from the upper boom
bracket, as shown at mid-right:
9
Use two /16" wrenches to remove
the hardware that secures top
of pump to upper boom bracket.
This will free the pump to rotate
downward.
existing pump
detach existing pump
from upper boom
detach existing pump
from lower mast
3. Remove the bottom of the exist-
ing pump from the lower mast
bracket, as shown at far right. Use
9
two /16" wrenches to remove the
hardware that secures bottom of
pump to lower mast bracket. This
will free the pump — take care
that it does not fall.
1031FP-INS-LAB-RevA11
16
Download from Www.Somanuals.com. All Manuals Search And Download.
mast
Installation of New Pump
Info: The hardware for attaching the Hydrau-
lic Pump (GF #1010S005) to the mast and
boom comes with the lift frame.
1. Attach the bottom of the new hydraulic pump
to the lower mast bracket with the pressure
knob and pump lever toward the bottom of
the pump, as shown at near right, and the
“FRONT” label toward you, as shown at far
right. Use a hex bolt, washers, and hex nut to
9
attach new pump
to lower mast
detail — attach new
pump to lower mast
secure pump to mast. Use two /16" wrenches
to tighten.
boom
2. Attach the top of the hydraulic pump to the
upper boom bracket, as shown at right. Use
a hex bolt, washers, and hex nut to secure
upper boom
bracket
9
pump to boom. Use two /16" wrenches to
tighten.
WARNING: Ensure that the Hydraulic Pump
is installed as described. If the Hydraulic
Pump is not properly installed, personal
injury and damage to the Hydraulic Pump
could result.
attach new pump
to upper boom
1031FP-INS-LAB-RevA11
17
Download from Www.Somanuals.com. All Manuals Search And Download.
MAINTENANCE SCHEDULE
Info: Only qualified persons should service and repair your Lumex patient lift. Regular
maintenance of your patient lift is necessary to ensure continuing proper and safe opera-
tion. Read and observe the following recommended maintenance schedule:
Item
Inspect
initially
Inspect/adjust
daily
monthly every 6
months
Base and Casters
Ensure that base is level (all four wheels touch the floor)
Ensure that caster mounting bolts and axle bolts are tight
Ensure that casters swivel and roll freely
Apply a light grease to caster ball bearings
Base Adjustments
Ensure that base opens and closes smoothly
Ensure that base locks securely in three positions
Ensure that locking knob is in place and operating properly
Inspect all base hardware for wear
Mast
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Ensure that mast is fully engaged and locked into mast sleeve
Inspect mast for bends or deflections
Inspect all mast hardware for wear
Boom
Inspect boom for bends or deflections
Ensure that boom is centered between the base legs
Inspect all boom hardware for wear
Spreader Bar
X
X
X
X
X
X
X
X
X
X
X
Inspect hanger post for wear or damage
Inspect sling hook for wear or deflections
Inspect all spreader bar hardware for wear
Slings
X
X
X
X
X
Inspect slings and straps for wear or damage
Inspect metal/plastic parts for wear or damage
Lubrication with light oil
X
X
Lubricate all pivot points
X
X
Lubricate top and bottom of actuator mounting
Cleaning
Clean with mild non-abrasive detergent
X
Pump
Ensure that pump mounting hardware is tight
Inspect pump mounting hardware for wear
X
X
X
1031FP-INS-LAB-RevA11
18
Download from Www.Somanuals.com. All Manuals Search And Download.
8 TROUBLESHOOTING
WARNING: If you experience a problem with your lift and are unable to service it your-
self, contact your Lumex dealer.
SYMPTOM
POSSIBLE CAUSE
Air in the cylinder
ACTION
Patient lift does not go up
or down
Bleed the cylinder by turning the release
knob and push down on the mast. Pump
up the lift again and turn the release knob
again.
Loss of hydraulic fluid (leaking
cylinder)
Replace any leaking cylinder
Patient lift does not go
down
Air in the cylinder
See above Directions
Info: If the hydraulic cylinder begins leaking fluid, do not operate. Contact your Lumex
dealer for a new cylinder.
1031FP-INS-LAB-RevA11
19
Download from Www.Somanuals.com. All Manuals Search And Download.
9 SPECIFICATIONS *
Info: All dimensions are given in inches unless otherwise specified.
Base width
Open: 39.7"
Closed: 25.4"
Base height Clearance
Base length
4.5"
51"
Lifting height range
Minimum: 24"
Maximum: 75.5"
Info: As measured from point where sling attaches
to spreader bar.
Range: 50.3"
Mast height
55.1"
Weight capacity
Caster size
400 lb (181 kg)
4.00" diameter, both locking and non-locking
Product Weight
Shipping weight
90 lb
103 lb
* Specifications are subject to change without notice.
1031FP-INS-LAB-RevA11
20
Download from Www.Somanuals.com. All Manuals Search And Download.
10 LIMITED WARRANTY
GF Health Products Inc. warrants this product to be free from manufacturing defects in material and
workmanship for its period of warranty that is in accordance with industry standards. This warranty
is extended only to the original purchaser/consumer or dealer/non-consumer of this new product
and to no other purchaser or transferee.
GF Health Products, Inc. warrants the Lumex® LF1031FP Hydraulic Patient Lift and its components to
be free from defects in workmanship and materials as listed below:
Frame / Parts: 1 Year
Hydraulic Pump: 1 Year
The Warranty period for the consumer commences on the first date a product is delivered to consumer
by seller/dealer. If the product is rented or leased, the warranty period commences on the invoice
date from GF Health Products, Inc. A copy of the invoice showing date of purchase must be provided
when submitting warranty claims. When proof of purchase date is not provided, warranty coverage
shall commence upon GF Health Products, Inc.’s invoice date to the dealer/purchaser.
If within the warranty period, the product or component part is proven to GF Health Products, Inc.’s
satisfaction to be defective, GF Health Products, Inc. shall provide, at its option, one of the following:
(1) repair or replacement of any defective or nonconforming part or product or (2) a credit and/
or refund of the original selling price. GF HEALTH PRODUCTS, INC.’S SOLE OBLIGATION AND YOUR
EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO SUCH REPAIR, REPLACEMENT,
CREDIT AND/OR REFUND. This warranty does not include any labor charges incurred in replacement
part(s) installation or any associated freight or shipping charges to the manufacturer.
For warranty service, please contact the authorized distributor from whom you acquired your GF
Health Products, Inc. product. Upon receiving notice of an alleged defect in a product, GF Health Prod-
ucts, Inc. will issue a return authorization. The defective product or part(s) must then be returned,
at the purchaser’s cost, for warranty inspection using the serial number as identification (or, if the
product is not serialized, lot number and date code) within thirty (30) days of return authorization
issue date. In the event you do not receive satisfactory warranty service, please contact GF Health
Products, Inc. at the address below. DO NOT return products to our factory without prior authorization.
LIMITATIONS AND EXCLUSIONS: The foregoing warranty excludes consumables, such as filters and
tubing, as these require routine maintenance. The foregoing warranty shall not apply to serial num-
bered products if the serial number has been removed or defaced. Products subjected to negligence,
abuse, misuse, improper operation, improper maintenance, improper cleaning, improper storage, or
damages beyond GF Health Products, Inc.’s control are not covered by this warranty, and that evalua-
tion shall be solely determined by GF Health Products, Inc. This warranty shall not apply to problems
arising from normal wear and tear or failure to follow instructions. The warranty shall also not apply
to products modified without GF Health Products, Inc.’s express written consent; nor shall it apply if
parts not manufactured by GF Health Products, Inc., or if parts not complying with original equipment
specifications are added to GF Health Products, Inc. products, or if the product or part is serviced by
an entity not authorized by GF Health Products, Inc.
THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WAR-
RANTIES AND IMPLIED WARRANTIES, INCLUDING BUT NOT LIMITED TO THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, AND
SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED
HEREIN, AND THE REMEDY FOR VIOLATIONS OF ANY IMPLIED WARRANTY SHALL BE
LIMITED TO THE REPAIR, REPLACEMENT, CREDIT AND/OR REFUND OF THE DEFEC-
1031FP-INS-LAB-RevA11
21
Download from Www.Somanuals.com. All Manuals Search And Download.
TIVE PRODUCT OR PART PURSUANT TO THE TERMS CONTAINED HEREIN. GF HEALTH
PRODUCTS, INC. SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL
DAMAGES WHATSOEVER.
This warranty gives you specific legal rights and you may also have other legal rights which vary
from state to state (province to province). Some states (provinces) do not allow the exclusion or
limitation of incidental or consequential damage, or limitation on how long an implied warranty lasts,
so the above exclusion and limitations may not apply to you.
The warranties contained herein contain all the representations and warranties with respect to the
subject matter of this document, and supersede all prior negotiations, agreements and understand-
ings with respect thereto. The recipient of this document hereby acknowledges and represents that
it has not relied on any representation, assertion, guarantee, warranty, collateral contract or other
assurance, except those set out in this document.
GF Health Products, Inc. (“Graham-Field”)
2935 Northeast Parkway
Atlanta, GA 30360
Tel 800-347-5678, Fax 800-726-0601
1031FP-INS-LAB-RevA11
22
Download from Www.Somanuals.com. All Manuals Search And Download.
11 INDEX
T
A
Transferring the patient from a bed to a wheelchair or
commode chair 13
Assembly instructions 8
B
Transferring the patient from a chair to a wheelchair or
commode chair 12
Before each use 14
Troubleshooting 19
C
Care and maintenance 14
U
Care of slings 15
Unpacking 7
CAUTION statement, significance 3
Checking for damage and wear 15
Cleaning 15
V
Components 8
W
Warnings 4
F
WARNING statement, significance 3
Warranty, limited 21
Features 6
Weight capacity 4
H
Handling Procedures 7
Hydraulic Pump 1010S005, Installation 16
I
Info statement, significance 3
Inspection 7
Introduction 3
L
Lifting the patient 11
Lifting the patient from a bed 11
Lifting the patient from the floor 12
Lowering the patient onto a bed 11
Lowering the patient to the floor 12
Lubrication 14
M
Maintenance schedule 18
N
NOTICE statement, significance 3
O
Operating Instructions 11
R
Removing the sling 13
S
Safety precautions 4
Safety statements, significance of 3
Shipping and transportation instructions 7
Specifications 20
Storage 7
1031FP-INS-LAB-RevA11
23