GE Water Dispenser GE OSMONICS MILLENIUMTM REVERSE OSMOSIS UNIT User Manual

GE OSMONICS  
MILLENIUM™ REVERSE OSMOSIS UNIT  
Operation  
and Maintenance Manual  
SUPERCEDES P/N 14693 REV. D  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MILLENIUM REVERSE OSMOSIS UNIT  
Installation, Operation and Maintenance Manual  
TABLE OF CONTENTS  
Page  
1-1  
CHAPTER ONE:  
GENERAL INFORMATION  
Section  
1.1  
1.2  
1.3  
1.4  
1.5  
1.6  
1.7  
1.8  
1.9  
The Manual  
Safety Summary  
Applications  
Contraindications  
Environmental Considerations  
Theory of Operation  
Fluid Components  
Monitors and Controls  
Optional Equipment  
1-3  
1-3  
1-4  
1-4  
1-5  
1-6  
1-7  
1-9  
1-11  
1-12  
1.10  
Specifications  
1.10.1  
1.10.2  
1.10.3  
1.10.4  
1.10.5  
1.10.6  
1.11  
1.12  
1.13  
1.14  
1.15  
Inlet Water Requirements  
Flow Rates  
Electrical Requirements  
Dimensions  
Membrane Performance Characteristics  
Environmental Requirements  
Disposal  
Electromagnetic Interference  
Service Assistance  
1-13  
1-14  
1-14  
1-15  
1-16  
Return Goods Authorization (RGA)  
Symbols and Abbreviations  
CHAPTER TWO:  
INSTALLATION GUIDELINES AND INITIAL  
SYSTEM START UP  
2-1  
Section  
2.1  
2.2  
2.3  
Installation Considerations  
Installation Procedure  
Product Water Analysis Procedure  
2-3  
2-3  
2-6  
CHAPTER THREE:  
SYSTEM OPERATION  
3-1  
Section  
3.1  
3.2  
Daily Start-up Procedure  
Shut Down Procedure  
3-3  
3-3  
14693e  
A
31 March 2003  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CHAPTER FOUR:  
ROUTINE MAINTENANCE  
4-1  
Section  
4.1  
Routine Maintenance of Millenium™ RO System  
Daily / Per Treatment Maintenance  
Weekly Maintenance  
Monthly Maintenance  
Quarterly Maintenance  
Semi-Annual Maintenance  
Annual Maintenance  
Specific Maintenance Procedures and Instructions 4-6  
4-3  
4-3  
4-4  
4-4  
4-4  
4-5  
4-5  
4.1.1  
4.1.2  
4.1.3  
4.1.4  
4.1.5  
4.1.6  
4.2  
4.2.1  
4.2.2  
4.2.3  
4.3  
4.3.1  
4.3.1.1  
4.3.1.2  
4.3.1.3  
4.3.2  
4.3.2.1  
4.3.2.2  
4.3.2.3  
4.4  
Motor Lubricant Procedure  
Leak Repair  
Product Recirculation Check Valve Test  
Millenium™ Cleaning Indications  
Pre Cleaning Preparation  
Pre Cleaning Procedure  
Cleaner Recirculation and Dwell Time  
Post Cleaning Rinse Procedure  
BIOSANPre Cleaning Preparation  
BIOSANCleaning Procedure  
BIOSANRecirculation and Dwell Time  
BIOSANPost Cleaning Rinse Procedure  
MilleniumDisinfection Indications  
Formaldehyde Disinfection Preparation  
Formaldehyde Pre Disinfection Procedure  
Formaldehyde Disinfection and Dwell Time  
Formaldehyde Post Disinfection Rinse Procedure  
“RENALIN®” Disinfection Procedure  
(Peracetic Acid and Peroxide-Based Compounds  
such as Renalin®, Minncare, & Peracidin)  
“RENALIN®” Pre Disinfection Preparation  
“RENALIN®” Disinfection and Dwell Time  
“RENALIN®” Post Disinfection Rinse Procedure  
Product Water Culture  
4-6  
4-6  
4-9  
4-9  
4-10  
4-10  
4-11  
4-12  
4-14  
4-14  
4-15  
4-16  
4-17  
4-17  
4-18  
4-18  
4-19  
4-21  
4.4.1  
4.4.1.1  
4.4.1.2  
4.4.1.3  
4.4.2  
4.4.2.1  
4.4.2.2  
4.4.2.3  
4.5  
4.5.1  
4.6  
4-22  
4-23  
4-23  
4-25  
4-26  
4-27  
4-27  
4-28  
Product Water Culture Procedure  
System Storage  
Formaldehyde Storage Procedure  
BIOSANStorage Procedure  
4.6.1  
4.6.2  
14693e  
B
31 March 2003  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CHAPTER FIVE:  
TROUBLESHOOTING  
5-1  
Table of Content  
Unit Fails to Run  
Unit Runs Intermittently  
Motor Runs Hot  
Water Produced While Unit is Off  
Low Pump Pressure  
Pressure Gauge Vibrates Erratically  
Excess Product Pressure (too high)  
Product Flow Too Low  
Product Flow Too High  
Leaks  
Low Percent Rejection  
TDS Value High  
RO Fails to Draw Disinfectant  
RO Pump Leaks  
5-3  
5-5  
5-5  
5-6  
5-6  
5-7  
5-7  
5-7  
5-8  
5-8  
5-9  
5-9  
5-10  
5-11  
5-11  
5-11  
Section  
5.1  
5.2  
5.3  
5.4  
5.5  
5.6  
5.7  
5.8  
5.9  
5.10  
5.11  
5.12  
5.13  
5.14  
5.15  
TDS Value Fluctuates  
CHAPTER SIX:  
DRAWINGS  
6-1  
7-1  
CHAPTER SEVEN:  
SPARE PARTS LIST  
14693e  
C
31 March 2003  
Download from Www.Somanuals.com. All Manuals Search And Download.  
This page intentionally left blank  
14693e  
D
31 March 2003  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO System  
CHAPTER ONE:  
GENERAL INFORMATION  
14693e – 31 Mar 03  
1 - 1  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e – 31 Mar 03  
1 - 2  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
1.1  
The Manual  
This manual has been prepared to provide the operator with information and instructions  
regarding the installation, use, maintenance and troubleshooting of the GE Osmonics  
Millenium™ Reverse Osmosis System.  
CAUTION:  
When used as a medical device, Federal law restricts this device to sale by or on the  
order of a physician. Per CFR 801.109 (b)(1).  
The manual has been written in narrative form supplemented with schematics and  
drawings for clarification. The operator can perform most procedures mentioned in this  
manual. Any exceptions will be clearly identified by a qualifying statement.  
1.2 Safety Summary  
Words in BOLD CAPITAL letters are used to identify labels on the device and key  
safety or qualifying statements. A list of all symbols and abbreviations is located at the  
end of this chapter. See Technote 120 for a glossary of relevant terms.  
This safety summary does not contain all of the safety statements in the manual. Other  
safety statements are included within the manual text and are enhanced and defined as  
follows:  
NOTE:  
Statements that provide further clarification.  
CAUTION:  
Statements identifying conditions or practices that could result in equipment or  
other property damage.  
WARNING:  
Statements identifying conditions or practices that could result in personal injury or  
loss of life.  
READ THIS MANUAL:  
Prior to operating or servicing this device, this manual must be read and understood.  
Keep this and other associated manuals for future reference and for new operators or  
qualified service personnel. A note sheet is provided at the end of each chapter for  
operators to make notations that may be valuable to other users.  
USE PROPER POWER CONNECTIONS:  
Use proper wiring and connection methods to satisfy hospital electrical codes.  
DO NOT REMOVE COVERS OR PANELS:  
To avoid electrical shock hazard, do not remove covers or panels when power is supplied  
to the device. Do not operate the device when covers or panels are removed.  
14693e – 31 Mar 03  
1 - 3  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
SHOCK HAZARD:  
Connect this device to a proper ground connection in accordance with the National  
Electrical Code. DO NOT under any circumstances remove the ground wire or ground  
prong from any power plug. DO NOT use an extension cord with this equipment. Keep  
the surface under the RO(s) free of water.  
DEVICE LABELING:  
Do not under any circumstances; remove any Caution, Warning or any other descriptive  
labels from the devices until the conditions warranting the label are eliminated.  
DO NOT OPERATE IN A FLAMMABLE ATMOSPHERE:  
To avoid fire or explosion, do not operate this device in an explosive environment or near  
flammable anesthetics.  
1.3 Applications  
The Millenium™ Reverse Osmosis (RO) System is designed to purify water by forcing  
water through a semi-permeable RO membrane. Water purified by reverse osmosis has  
had approximately 95% of the dissolved ions and approximately 99% of most other  
contaminants removed. The quality of the purified water, referred to as product water or  
permeate, depends on the quality of the feed water and normally is equal to or exceeds  
AAMI standards. Such water is suitable for hemodialysis applications.  
1.4 Contraindications  
WARNING:  
GE Osmonics systems are not designed, sold, or intended for use in producing  
water for injection  
Furthermore, GE Osmonics systems are not intended to be used outside of the device  
specifications and limitations, as outlined in this manual and other related materials.  
14693e – 31 Mar 03  
1 - 4  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
1.5 Environmental Considerations  
Prior to the installation of the Millenium™ RO System, it will be necessary to provide  
utilities and create an environment suitable for the trouble free operation of the RO  
system and its accessories.  
POWER:  
The Millenium™ operates on 115 VAC single-phase power. Histories of power  
failure, power surges, and low line voltages should be noted and reported to the  
manufacturer or their agent as they may create adverse conditions for the operation  
of equipment.  
WATER:  
Reverse osmosis systems require a reliable water source. The Millenium™ RO  
needs a minimum of 30-PSI pressure at a minimum flow of 2 gallons per minute. It  
is important that water is flowing at the design flow rate when testing water  
pressure rather than reading static pressure.  
DRAIN:  
A drain outlet is required by the RO system. The drain must have a minimum  
capacity of 2 gallons per RO per minute of continuous flow for as long as and as  
often as the RO is operating.  
TEMPERATURE:  
The Millenium™ RO is normally connected to a cold water supply. The  
performance specifications of the RO are based on 77°F (25°C) feed temperature.  
Each degree Fahrenheit the feed water temperature falls, the RO product flow  
decreases approximately 1.5%, and each degree Celsius drop, the product flow is  
reduced by 3% (See Technote 113 for the “Temperature Correction Factors” chart).  
In climates where the winter water temperature drops significantly below 77°F it  
may be necessary to temper (heat) the water.  
WARNING:  
The temperature of the tempered water should be monitored closely with a  
thermometer as high temperatures can damage the RO membranes and possibly  
harm the patient. Refer to the inlet water requirements (Section 1.10.1) for the  
maximum operating temperature for the Millenium™ RO.  
CAUTION:  
It is necessary to keep the RO from freezing to prevent serious damage to the RO  
unit.  
NOISE:  
The RO unit does not generate significant noise levels. Hard walls may reflect  
noise and will make the RO seem louder.  
14693e – 31 Mar 03  
1 - 5  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
WATER PURITY:  
Plumbing materials can significantly contribute to the contamination of the water.  
Metallic pipes (with the exception of 316 stainless steel) must be eliminated once  
the purification process has begun. Schedule 40 or 80 PVC pipes, polypropylene,  
PVDF, and other FDA recommended materials are suitable for most grades of  
purified water. Care must also be exercised in the choice of a thread sealant.  
Teflon tape is suitable for all threaded connections in this system.  
WARNING:  
Ordinary pipe dope must be avoided since it may leach objectionable and  
potentially dangerous impurities into the water.  
1.6  
Theory of Operation  
The Millenium™ Portable Reverse Osmosis system (RO) is a fully cabinetized unit that  
purifies a given feed water (by way of reverse osmosis) for use in hemodialysis  
applications. The Millenium™ RO possesses monitoring for pressure, product water  
quality and membrane performance (percent rejection). The RO pump provides the  
pressure required to push water through the RO membranes. The fluid controls provide a  
means of managing flow rates and pressures. The following paragraphs detail the  
function of these fluid control components, which are used within the RO.  
14693e – 31 Mar 03  
1 - 6  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
1.7 Fluid Components  
Inlet Solenoid:  
Located inside the Millenium™ RO at the inlet, and serves to shut off the water  
supply to the system, until it is actuated when the Inlet water valve switch is placed  
in the ON position.  
Pressure Adjust Regulator:  
Used to adjust the product pressure. The pressure is displayed on the product  
pressure gauge, on the front of the RO.  
Inlet Water Check Valve:  
Prevents back-flow of disinfectants or cleaning agents.  
Inlet Water Sensor:  
Monitors the quality of the inlet water. Inlet quality is compared to the outlet  
(product) water quality and is displayed as "percent rejection" on the front panel.  
High-Pressure Pump:  
Provides up to a 200 PSI boost of the inlet water to the Reverse Osmosis Membrane  
Assembly (ROMA). The RO pump pressure is displayed on a 0 to 300 PSI gauge  
mounted outside the RO, above the product pressure gauge. The pressure is factory  
set at 190 PSI.  
Reverse Osmosis Membrane Assemblies (ROMAs):  
Two are included in the Millenium™ RO. A ROMA is an assembly that includes  
the membrane housing, end caps, fittings, and a membrane. Depending on the  
model, thin film membrane ROs can either provide up to 500 or 750 gallons per day  
of product water, depending on membrane type, condition, and water temperature.  
Waste Flow:  
The Waste Flow is controlled by a length of 1/8" tubing. The length of the tube is  
factory determined to correctly govern the wastewater flow to drain.  
Operate/Flush Valve:  
The Operate/Flush Valve on the front panel is used to flush the membrane surface  
of rejected contaminants, and to accelerate the rinse out of cleaners or disinfectants  
from the RO. Some pump noise (chattering) is normal in the flush position. The RO  
is in "flush" from the full open (flush) to just before the fully closed (operate)  
position.  
CAUTION:  
Excessive “chattering” may be eliminated by slowly moving the valve from the full  
flush position towards “OPERATE”.  
14693e – 31 Mar 03  
1 - 7  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Product Water Pressure:  
Product Water Pressure is read on a 0 to 100 PSI gauge on the front panel. Product  
pressures will register only when the product is back pressured (product hose  
removed from fitting or connected to dialysis unit).  
Product Recirculation Check Valve:  
Prevents over-pressuring of the product supply. When the product pressure exceeds  
the set point of the pressure regulator, the product recirculation check valve opens.  
Excess product water is then recirculated back to the inlet water flow.  
Product Water Sensor:  
The Product Water Sensor is used in conjunction with the inlet water sensor by the  
percent-rejection circuitry to calculate the membrane rejection rate. The rate of  
rejection is displayed from 70% - 99% in a 5 LED array on the front panel.  
Product Check Valve:  
The Product Check Valve is used to prevent anything from flowing backward into  
the product side of the RO membranes.  
CAUTION:  
Back-flow pressure into the product line can cause damage to the membranes.  
Total Dissolved Solids (TDS) Sensor:  
Samples the product water quality via the sensor probe, and the final quality of the  
product water is indicated on a digital display on the front panel as TDS in mg/L  
and ppm. The monitor is temperature compensated to adjust the display value to  
changing water temperature.  
Disinfect Inlet Port:  
The Disinfect Inlet Port is a quick-disconnect valve that allows access to the inlet  
water side of the pump. The port is used when cleaners or disinfectants are  
introduced to the RO. The valve can also be used to obtain feed water samples for  
testing feed water for chlorine/chloramine, pH and hardness levels.  
WARNING:  
Water from this port is NOT RO product water and should not be used for dialysis  
solutions.  
Disinfect Switch:  
The Disinfect Switch is used to introduce disinfectants and cleaners into the RO.  
For safety, this switch must be held continuously to operate.  
CAUTION:  
When drawing disinfectants or cleaning solutions, do not allow the RO to draw air.  
This may cause irreparable damage to the RO pump.  
14693e – 31 Mar 03  
1 - 8  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
1.8 Monitors and Controls  
Inlet Water Valve Switch:  
The Inlet Water Valve Switch is mounted on the control panel. The switch is used  
to energize a solenoid, pressurizing the system with feed water.  
RO Pump Power Switch:  
The RO Pump Power Switch is mounted on the control panel. The switch is used to  
energize the RO pump/motor. The inlet water switch must be ON for the  
pump/motor to operate.  
Disinfect Switch:  
The Disinfect Switch is mounted on the front panel. This switch is used to  
introduce disinfectant or cleaning agents into the RO through the disinfect port.  
Pressure Adjust Regulator:  
The Pressure Adjust Regulator inside the RO is used to set the maximum product  
pressure needed to supply a dialysis machine. Adjusting this regulator will also  
result in small adjustments in the pump pressure.  
Product Pressure Gauge:  
The Product Pressure Gauge reads the product pressure from 0 to 100 PSI. Product  
pressure will read zero when the product line is not connected to a dialysis unit and  
the product hose is connected to the product fitting and the valve is open.  
Pump Pressure Gauge:  
The Pump Pressure Gauge reads the pump pressure from 0 to 300 PSI. It is used  
for monitoring during operation and when adjusting the pump pressure, and is  
located on the front of the RO.  
Operate/Flush Valve:  
The Operate/Flush Valve is a 1/4-turn ball valve. When in the operate mode, the  
valve is in the closed position. When the valve is turned at any point towards flush,  
the valve is open. The closer to the flush point, the more the valve is open. When in  
flush, the valve bypasses the waste flow tube.  
Disinfect Port:  
The Disinfect Port is a quick-disconnection port for the introduction of disinfectant  
or cleaning agents. The port has a check valve that closes when there is no hose  
connected to the fitting.  
Feed Port:  
The Feed Port is a quick-disconnection port for the introduction of feed water to the  
RO. The port has a check valve that closes when there is no hose connected to the  
fitting.  
14693e – 31 Mar 03  
1 - 9  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Product Port:  
The Product Port is a quick-disconnection port for the output of purified product  
water. The port has a check valve that closes when there is no hose connected to  
the fitting.  
Waste Port:  
The Waste Port is a quick-disconnection port for the output of concentrate or waste  
to the drain. The port has a check valve that closes when there is no hose connected  
to the fitting.  
Percent Rejection Display:  
The Percent Rejection Display is a 5 LED array that indicates how the RO  
membranes are performing. The lights correspond to 99%, 95%, 90%, 85%, and  
>70% rejection. The 99% light, when illuminated, indicates a need to evaluate the  
RO system performance. RO units do not normally perform in excess of 99%;  
verify this reading with an independent meter and take corrective action if  
indicated. Some RO membranes will achieve this level of performance under  
optimum conditions.  
Product Water Total Dissolved Solids (TDS):  
The Product Water Total Dissolved Solids (TDS) digital display indicates product  
water purity in mg/L (ppm). It is temperature compensated to adjust the value of the  
display with changing water temperature for better accuracy.  
Less Than 0.5 mg/L Check System Lamp:  
The Less Than 0.5 mg/L Check System lamp illuminates when the TDS reading is  
0.5 mg/L or less. This indicates a need to evaluate the RO system performance.  
RO units do not normally have this low product TDS. Verify this reading with an  
independent meter and take corrective action if indicated. RO membranes may  
achieve this level of performance under optimum conditions (good feed water).  
Alarm Set:  
The Alarm Set switch allows the operator to set the TDS levels at which the alarm  
will sound. The alarm should be set at twice the initial TDS reading. If the set  
point is exceeded, an audible alarm will sound.  
Alarm Mute:  
The Alarm Mute switch will mute the audible alarm for 30 - 45 seconds. A red  
indicator lamp is displayed on the switch when the mute switch has been pressed.  
14693e – 31 Mar 03  
1 - 10  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
1.9 Optional Equipment  
GE Osmonics offers a number of optional items to adapt the equipment to meet specific  
needs. Options are available to improve portability, and to improve the quality of the  
feedwater to the RO. Frequently requested options are briefly described in the following  
paragraphs. If more information is needed, or if other options are desired, please consult  
the factory.  
Pretreatment Cart holds the Millenium™ RO and feedwater treatment equipment. The  
cart has a bracket that provides for the attachment of prefilter cartridge housings. The  
cart will hold any combination of two or three of the following tank sizes, .25 c.f., .47  
c.f., or .60 c.f..  
Carbon Filter Tanks remove chlorine/chloramine from the feed water, preventing  
damage to the RO membrane and injury to patients. The filter incorporates a bed of  
granular activated carbon. Three sizes of carbon tanks come in semi-automatic control  
valve and exchange head styles.  
WARNING:  
Carbon is not regenerated and must be replaced when exhausted.  
WARNING:  
A suitable carbon filter should always be present as pretreatment for the RO unit in  
order to prevent chlorine/chloramine damage to the thin film RO membrane and  
harm to the patient. Osmonics strongly recommends the use of two carbon tanks,  
used in a series configuration.  
Softeners remove calcium, magnesium, and other scale producing contaminants from the  
feed water. Their removal prevents the build-up of scale on the membrane surface. A  
control valve drives the regeneration of the softener, and will remove the hardness ions  
with a brine solution and rinse the ion exchange bed afterwards. The control valve is  
normally governed by a semi-automatic time clock. There are three sizes of softener  
tanks available.  
Dealkalizers are used to lower the pH of alkaline feed water. Water with pH levels  
above 8.5 with chloramines present can reduce the performance of the carbon filters and  
thin-film membranes. A control valve drives the regeneration rinsing of the dealkalizer.  
The control valve is normally governed by a semi-automatic time clock. There are three  
sizes of dealkalizer tanks available.  
14693e – 31 Mar 03  
1 - 11  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
1.10 Specifications:  
1.10.1 Input Water Requirements  
Minimum  
Maximum  
N/A  
Input flow Rate  
Temperature (water)  
NOTE:  
2.0 GPM  
1.7°C (35°F)  
32°C (90°F)  
Operating at lower temperatures may make it impossible for the Millenium™ to  
produce adequate permeate for the dialysis machine.  
pH (operating)  
TF Membranes  
4.0  
11.0  
pH w/chloramines present (operating)  
TF Membranes  
4.0  
8.5  
pH (cleaning)  
2.0  
11.0  
TF Membranes  
Pressure, dynamic  
30 PSI  
100 PSI  
1.10.2 Flow Rates  
Product Flow based on the following conditions:  
New Membranes  
Operating Pressure:  
Product Pressure:  
190 PSI  
1 PSI  
Feed Water Temperature:  
77°F (25°C)  
GPM LPM GPD LPD  
.34  
.52  
1.31 500  
1.97 750  
1892.5  
2839.1  
Membrane flux may vary ±15%  
Waste Flow: .84 GPM (3.2 LPM) (±10%)  
1.10.3 Electrical Requirements  
Motor  
Voltage  
Hertz  
1/3 HP  
115 VAC / 230 VAC  
60 / 50  
8 / 5  
Amps  
Phase  
single  
14693e – 31 Mar 03  
1 - 12  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
1.10.4 Dimensions  
Vertical Cabinet  
Horizontal Cabinet  
8 inches  
Height  
25 ½ inches  
15 inches  
8 inches  
Depth  
15 inches  
Width  
25½ inches  
Weight – Operating – 55 lbs.  
1.10.5 Membrane Performance Characteristics  
Salt rejection (TF Membranes):  
95% minimum at initial testing at factory.  
Test Solution:  
Conductivity:  
Operating Pressure:  
Product Back pressure:  
Inlet water 1000 PPM NaCl  
2000 micro mhos-cm  
190 PSI, ±10 PSI  
0 PSI  
1.10.6 Environmental Requirement  
Minimum  
Maximum  
Ambient Temperature  
Storage Temperature  
Altitude  
4°C (39°F)  
32°C (90°F)  
2°C (36°F)  
N/A  
32°C (90°F)  
10,000 feet  
CAUTION:  
This equipment must not be allowed to freeze. Irreparable harm to various  
components, including RO membranes may result.  
NOTE:  
The user/operator should recognize that moisture can be caused by condensation  
and is not necessarily an equipment leak. This equipment will function in the  
presence of condensation.  
1.11  
Disposal  
Disposal of this product or parts must be carried out according with local disposal codes.  
14693e – 31 Mar 03  
1 - 13  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
1.12  
Electromagnetic Interference  
This equipment can generate, uses and can radiate radio frequency energy and, if not  
installed and used in accordance with these instructions, may cause harmful interference  
to other devices in the vicinity. However, there is no guarantee that interference will not  
occur in a particular installation. If this equipment does cause harmful interference with  
other devices, which can be determined by turning the equipment off and on, the user is  
encouraged to try to correct the interference by one or more of the following measures:  
Reorient or relocate the receiving device  
Increase the separation between the equipment  
Connect the equipment into an outlet on a circuit different from that to  
which the other device(s) is connected  
Consult the field service technician or manufacturer for help.  
1.13 Service Assistance  
If service assistance is required, please take the following steps:  
Consult the “Troubleshooting” section of this manual (Chapter Five). If the problem  
cannot be identified and corrected by any of the procedures found in that section, then...  
Contact your Facility Equipment Technician. If the technician is unable to help then…  
Call GE Osmonics Technical Service Department at (800) 633-3080 or (206) 395-2200.  
Technicians are available for all calls between 7:00 a.m. and 4:30 p.m. PST, Monday  
through Friday. Technicians are also available at other times for emergency calls only.  
Product consultants will be on hand to discuss the problem with you and endeavor to  
rectify it over the phone. If the problem appears to be of a more serious nature, you will  
be given instructions regarding the action to be taken. Prior to making the phone call,  
you must be prepared to answer two questions:  
1. What RO do you have, i.e. Millenium™  
2. What is the serial number of your RO? (Label, found on back of RO.)  
RO SERIAL NUMBER: ______________________  
14693e – 31 Mar 03  
1 - 14  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
1.14 Return Goods Authorization (RGA) Procedure  
If you wish to return goods for, warranty evaluation and/or credit please have your  
original sales order, invoice and device serial number available when you call GE  
Osmonics. Call GE Osmonics at (800)-633-3080 and request Technical Service. A GE  
Osmonics Technical Service representative will provide instructions and a return  
authorization number, which needs to be clearly written on the outside of the box used to  
ship your materials. All equipment must be shipped to GE Osmonics with the freight  
prepaid by the customer. Call our Customer Service Center with any questions or issues  
concerning freight claims and a representative will discuss your situation.  
All materials to be returned must be rendered into a non-hazardous condition prior to  
shipping.  
14693e – 31 Mar 03  
1 - 15  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
1.15 Symbols and Abbreviations  
“RENALIN®”  
Refers to Peracetic Acid and Peroxide-based Compounds, such as  
Renalin®, Minncare™, Peracidin™ (20% peroxide, 4% Peracetic  
Acid)  
ASTM  
AAMI  
CA  
American Society for Testing and Materials  
Association for the Advancement of Medical Instrumentation  
Cellulose Acetate  
C
Celsius  
CSA  
cc  
Canadian Standards Association  
Cubic Centimeters  
cf  
Cubic Foot (feet)  
cfu/ml.  
cm  
Colony Forming Units per milliliter  
Centimeters  
DI  
Deionization  
F
Fahrenheit  
FDA  
Ft  
Food and Drug Administration  
Foot (feet)  
GPD  
GPM  
GHT  
gr  
Gallons Per Day  
Gallons Per Minute  
Garden Hose Thread  
Grains  
GAC  
HZ  
Granular Activated Carbon  
Hertz  
HP  
Horse Power  
lbs.  
Pounds  
LED  
LPD  
LPM  
m.  
Light Emitting Diode  
Liters Per Day  
Liters Per Minute  
Meter  
mg/L.  
ml.  
Milligrams Per Liter  
Milliliters  
14693e – 31 Mar 03  
1 - 16  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
MNPT  
MSDS  
Megohm/cm  
mw  
Male National Pipe Thread  
Material Safety Data Sheet  
Million Ohms/Centimeter  
Molecular Weight  
NaCl  
N/A  
Sodium Chloride  
Not Applicable  
NPT  
NSF  
National Pipe Thread  
National Sanitation Foundation  
Polyamide  
PA  
P/N  
Part Number  
ppm  
psi  
Parts Per Million  
Pounds Per Square Inch  
Polyvinyl Chloride  
Quick Disconnect  
PVC  
QD  
RO  
Reverse Osmosis  
ROMA  
TB  
Reverse Osmosis Membrane Assembly  
Terminal Barrier Strip (Main): “TB1” (Example: “TB1-9” refers  
to the terminal marked number 9 on the Main Terminal Barrier  
Strip)  
TF  
Thin Film  
TDS  
UL  
Total Dissolved Solids  
Underwriters Laboratory  
Volts Alternating Current  
Water for Injection  
Weight  
VAC  
WFI  
WT  
~
Volts Alternating Current  
14693e – 31 Mar 03  
1 - 17  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
NOTES:  
14693e – 31 Mar 03  
1 - 18  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO System  
CHAPTER 2:  
INSTALLATION GUIDELINES AND INITIAL SYSTEM START-UP  
14693e - 31 Mar 03  
2 - 1  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
2 - 2  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
2.1  
Installation Considerations  
1.  
Select a location for the Millenium™ near the required utilities and dialysis  
equipment.  
2.  
3.  
The RO unit requires 115 VAC power for operation and consumes 8 amps of  
current (the 230 VAC model consumes 5 amps).  
The RO unit requires nominally 2 GPM water at 30 PSI. Warmer feed water (up  
to 25°) will produce more permeate, and will consume more water. The water  
may also require some pretreatment to ensure maximum performance and  
suitability. GE Osmonics recommends that feed water be chlorine/chloramine  
free and solids above 5 microns in size be filtered out.  
4.  
5.  
6.  
A drain with a minimum capacity of 2 GPM is required for the RO unit.  
Back flow prevention may be required in accordance with local codes.  
Special connections may be needed to connect to water supply. Connections vary  
with different facilities.  
2.2  
Installation Procedures:  
CAUTION:  
If a carbon tank is present on the RO system, the carbon tank must be completely  
rinsed of all carbon fines and ash before connecting to the RO unit. Otherwise,  
irreparable damage to the TF membrane will result (refer to Carbon Manual).  
WARNING:  
The Millenium™ RO is stored with 1 wt.% sodium metabisulfite from the factory.  
Follow all Material Safety Data Sheet (MSDS) guidelines that are included with the  
RO.  
NOTE:  
For initial start-up, the large removable panel must be removed to allow access to  
the pump and regulator for adjustments as well as to remove shipping materials.  
The panel is held on with twelve stainless steel screws. Use caution any time a  
panel is removed from the Millenium™.  
1.  
2.  
Connect the Millenium™ RO inlet tubing GHT (garden hose thread) connector to  
the inlet cold water supply, and the QD (quick-disconnect) end to the “Feed”  
connection on the front panel.  
Connect the waste water line to the waste QD connection on the front of the RO,  
and to the drain in accordance with the local plumbing code. Making this  
connection allows the waste flow by opening a check valve; disconnecting closes  
it.  
3.  
Connect the product line QD connector to the product fitting on the front panel.  
Connecting together the male and female fittings (hose to RO) opens a check  
14693e - 31 Mar 03  
2 - 3  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
valve allowing product flow. When the hose is disconnected, the product flow is  
stopped.  
4.  
Make sure that the inlet water valve switch and the RO pump power switch are in  
the OFF position. Plug the power cord into a 115 VAC (230 VAC) grounded  
outlet. Do not use an extension cord or an adapter.  
5.  
6.  
7.  
Turn the water supply to the RO on.  
Turn the operate/flush valve to the flush position.  
Turn the inlet water valve switch on. Water will start to fill the RO. Allow the  
RO to rinse to drain for five minutes after water begins to flow from the drain  
line. The pump pressure gauge should indicate some pressure.  
8.  
9.  
Turn the RO pump power switch on.  
NOTE:  
The operate/flush valve may need to be rotated part way towards the operate  
position if the pump makes excessive noise.  
Slowly turn the operate/flush valve to the operate position. Pump pressure should  
not exceed 200 PSI. If 200 PSI is exceeded, adjust pump pressure on pump head  
located inside the cabinet. See instructions in Figure 1 below:  
Figure 1  
14693e - 31 Mar 03  
2 - 4  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
10.  
Let the RO run for an additional 15 minutes or until the percent rejection and  
TDS values have stabilized.  
WARNING:  
Ensure that TDS and membrane performance (% rejection) are normal for the  
regions water quality. Consult your facility physician. An AAMI analysis and  
bacteria test of the product water must be obtained and interpreted before using the  
RO unit for hemodialysis treatment (refer to the Product Water Analysis procedure  
and the Product Water Culture Procedure).  
11.  
12.  
Set the product water TDS alarm limit switch as prescribed by the physician. The  
switch is normally set to twice the initial TDS reading (or “next higher” setting) at  
start-up.  
WARNING:  
The TDS alarm limit should not be moved from this setting unless instructed by a  
physician.  
After a minimum of 3 hours run time, disinfect and rinse the RO unit with an  
approved disinfectant before using for hemodialysis treatment (refer to the  
Formaldehyde and ‘RENALIN®’-type disinfection procedures in chapter four).  
CAUTION:  
New RO units with thin film membranes must be rinsed to drain with normal  
pressures for a minimum of three (3) hours before exposure to disinfectants.  
WARNING:  
After disinfection, the product water must be verified free of residual disinfectant.  
13.  
The RO product may be used in conjunction with hemodialysis treatment  
equipment only after obtaining test results which validate compliance with  
excepted standards.  
AAMI Product Water Panel  
Microbiological Culture (colony count) results  
Endotoxin (pyrogen) level testing (if applicable)  
14.  
The Millenium™ RO is now ready for use.  
14693e - 31 Mar 03  
2 - 5  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
2.3  
Product Water Analysis Procedure:  
In order to ensure that the RO is performing at or above AAMI (Association for the  
Advancement of Medical Instrumentation) Standards, it is necessary to perform an  
AAMI Water Quality Analysis test. For new units, the RO must be operated (rinsed to  
drain) for at least two hours before taking a sample, to ensure that the membranes are free  
of preservative or other contaminants which would invalidate the test.  
WARNING:  
There are many documented clinical risks in using less than AAMI Standard  
product water (refer to Technote 103, "AAMI Hemodialysis Water Quality  
Standards").  
NOTE:  
Follow the procedures recommended by your water quality-testing laboratory if  
they vary from the following steps.  
Materials Required: Produce water sampling kit, sterile gauze, tape, and container (such  
as a small bucket).  
1.  
Ensure that all pre-RO devices are operating correctly and are performance  
verified by the appropriate testing.  
2.  
3.  
Connect the RO unit to the required utilities (feed water, drain, power).  
Operate the RO unit with product and waste flows to drain for at least 15 minutes  
after the product water quality has stabilized before drawing the water sample.  
4.  
5.  
Hold the product line over the container and run product water for 60 seconds.  
Aseptically remove the lid from the sample bottle and rinse the bottle twice with  
RO product water.  
6.  
7.  
Fill the bottle completely with RO product water and cap it immediately.  
Fill out the laboratory requisition form. Ensure the requisition form has, at a  
minimum, information to identify the facility or location of the RO, the serial  
number or facility equipment number to ensure which RO was tested and name,  
phone number and address of whom to contact.  
8.  
Place the bottle and the completed requisition in the pre-addressed mailer and  
mail it directly to the laboratory.  
NOTE:  
Total chlorine must be measured on-site. Tests for Chlorine/Chloramine (total  
chlorine) are not typically reported on the AAMI Analysis. Chlorine and  
chloramine dissipate from the water sample over time and reported results will be  
inaccurate.  
NOTES:  
14693e - 31 Mar 03  
2 - 6  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
2 - 7  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO System  
CHAPTER 3:  
SYSTEM OPERATION  
14693e - 31 Mar 03  
3 - 1  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
3 - 2  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
3.1 Daily Start-up  
WARNING:  
Verification of the absence of all sanitizers, cleaners, disinfectants and  
chlorine/chloramines must be performed before RO product water is used for any  
patient related uses.  
1.  
2.  
3.  
Connect the RO unit to the required utilities (feed water, drain, and power).  
Ensure that all the valves in the water supply line to the RO are open.  
Ensure that all pre-RO devices have been flushed and tested and are operating  
within their specifications.  
4.  
5.  
6.  
Ensure that the product rinse line is connected to the RO, with the other end in a  
drain.  
Move the inlet water valve switch to the open position. When water flows to  
drain, the RO pump power switch can be moved to the on position.  
The "Percent Rejection" monitor should indicate 95% rejection within 2 minutes  
of start-up. It is normal for rejection to drop off until membranes have rinsed up  
to optimum quality.  
7.  
The TDS monitor should stabilize within 2 minutes of start-up. The TDS during  
this time may be higher than the alarm set point. The audible alarm may be  
muted at this time by pressing the alarm mute switch. The alarm mute switch will  
silence the audible alarm for 30 - 45 seconds each time the switch is used.  
WARNING:  
When the RO is used for dialysis service, a test for disinfectant residual MUST be  
performed.  
8.  
Record data on a daily log sheet.  
WARNING:  
Any data that does not meet minimum standards set at your facility must be  
reported to the medical director. The RO should not be placed in service until  
readings are acceptable.  
9.  
The Millenium™ RO is now ready for connection to the dialysis machine.  
Shut Down Procedure  
3.2  
1.  
Turn off the RO pump power switch.  
Turn off the inlet water valve switch.  
Turn off the inlet water supply.  
2.  
3.  
14693e - 31 Mar 03  
3 - 3  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.  
Briefly turn on the inlet water valve switch. This will release pressure from the  
feed water hose. When the pressure is released, turn the switch off.  
NOTE:  
If the Millenium™ RO is to be stored for more than 24 hours (RO not being  
operated), the RO should be stored in an approved disinfectant or storage solution  
(Not ‘RENALIN®’ or other type oxidizing products). Refer to the disinfection or  
storage procedure in this manual.  
5.  
Disconnect from the utilities and store the RO.  
14693e - 31 Mar 03  
3 - 4  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
NOTES:  
14693e - 31 Mar 03  
3 - 5  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
3 - 6  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO System  
CHAPTER FOUR:  
ROUTINE MAINTENANCE  
14693e - 31 Mar 03  
4- 1  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
4- 2  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.1  
Routine Maintenance of Millenium™ RO System  
GE Osmonics Millenium™ RO systems have been designed to operate with a minimum  
of operator attention. Like all mechanical systems they will operate longer and with less  
trouble when operator maintenance is performed regularly. Operator maintenance on a  
GE Osmonics reverse osmosis system is limited to maintaining performance logs,  
cleaning and disinfection, replacement of damaged or failed parts and periodic  
lubrication. The maintenance procedures have been categorized by their frequency of  
action and are as follows:  
NOTE:  
The first part of this section provides the timelines and simple instructions for  
maintenance of your RO. The specific instructions and procedures for the more  
complex maintenance items are provided in the second part of this section.  
WARNING:  
Prior to performing any electrical maintenance unplug the RO power cord from the  
wall receptacle.  
4.1.1 Daily Maintenance  
1.  
2.  
Test and record the hardness level of the feed water in the pretreatment log.  
Observe and record salt level in the brine tank (for systems with water softener  
pretreatment).  
3.  
Observe and record the Chlorine/Chloramine concentrations in the pretreatment  
log. Follow the instructions included with each test kit.  
WARNING:  
Take corrective action or contact your local service representative if any optional  
pretreatment device fails to provide adequate treatment of feed water before  
continuing with the use of the RO unit.  
4.  
If you are using cartridge pre-filters with your Millenium™, observe and record  
the pre-filter inlet and outlet pressure gauges on the operation log. The difference  
in the pressure values will help measure the condition of the filter cartridge.  
When the pressure differential (delta P) increases 8 PSI above the clean, initial  
delta P, the cartridges need to be changed (or sooner if the pre-filter core appears  
soiled). To change the pre-filter, the water supply must be turned off. The pre-  
filter inlet and outlet gauges should read zero PSI. The pre-filter bowl can now be  
removed by rotating it clockwise. Remove the old filter, install the new filter and  
re-install the filter bowl.  
5.  
Observe and record the percent rejection and TDS water quality readings in the  
operation log.  
14693e - 31 Mar 03  
4- 3  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.1.2 Weekly Maintenance  
1.  
Observe, or measure, and record the waste and product flow rates on the operation  
log. Allow the water temperature to stabilize before attempting to record the  
results. This will help prevent errors due to fluctuating temperature.  
2.  
Observe, or measure, and record the feed water temperature. Systems with  
blending valves have thermometers installed in the plumbing. A hand held  
thermometer may also be used. The water temperature has a significant effect in  
the product flow rate of the RO membrane and on the evaluation of the  
membranes’ need to be cleaned.  
3.  
Disinfect the system if needed. Refer to the disinfection procedure in this  
chapter.  
4.1.3 Monthly Maintenance  
1.  
Take a product water culture to test for bacteria colony forming units count (<200  
cfu/ml). Refer to the product water culture procedure in this chapter.  
2.  
Visually inspect the pre-filter core, if you are using cartridge pre-filtration.  
CAUTION:  
On occasion, the pressure differential between the two pre-filter gauges will not  
alter significantly, however, the pre-filter will be soiled. If the core of the pre-filter  
is dirty, particulates are breaking through the pre-filter and traveling downstream  
and causing damage to the pump and RO membranes. If this is the case, the pre-  
filter will require more frequent replacement on a routine basis.  
3.  
4.  
5.  
Check the pump pressure settings. The pressure should be 190 PSI ± 10 PSI, and  
should not exceed 200PSI.  
Check the maximum product pressure setting. If it is above the original set point,  
reduce the pressure with the regulator inside the cabinet.  
Perform trend analyses from the log sheets.  
4.1.4 Quarterly Maintenance  
1.  
2.  
3.  
4.  
Clean the RO Membranes. Refer to the RO membrane cleaning procedure in this  
chapter.  
Disinfect the RO machine. The RO should be disinfected after cleaning. Refer to  
the membrane disinfection procedure in this chapter.  
Verify the TDS and percent rejection readings with an independent monitoring  
device every 3 months.  
Check the quick-disconnect fittings and hoses. They should seat securely and  
fluid should shut off when they are disconnected.  
14693e - 31 Mar 03  
4- 4  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.1.5 Semi-annual Maintenance  
1.  
2.  
Check and clean the Millenium™ cooling fan and grill.  
Check the Product Recirculation check valve for proper operation. Refer to the  
check valve test instructions in this chapter.  
4.1.6  
Annual Maintenance  
1.  
An AAMI Water Quality Analysis of the product water should be conducted at  
least yearly. (Refer to the Water Quality Analysis Procedure in Chapter two.)  
2.  
3.  
4.  
5.  
6.  
7.  
8.  
Check the monitoring for accuracy.  
Check the gauges for correct functioning.  
Check the ball valves for correct operation.  
Check the tubing and fittings for leaks or wear.  
Check the switches and lights for correct operation.  
Check all pre RO devices for correct operation.  
Lubricate the motor. Refer to the lubricating instructions in this chapter.  
14693e - 31 Mar 03  
4- 5  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.2  
Specific Maintenance Procedures and Instructions  
4.2.1  
Motor Lubricant Procedure  
1.  
Remove the plastic plugs at both ends of the motor housing near the motor shaft.  
2.  
Add three drops of SAE 20 wt. non-detergent motor oil into the opening.  
Replace the plastic plugs.  
3.  
4.  
Turn on the motor and run for 10 minutes.  
Leak repair  
4.2.2  
General information: Any leak from any part of the system should be corrected as soon as  
possible. Left unattended, leaks can cause secondary damage necessitating expensive  
repairs.  
NOTE:  
Use FDA/NSF approved silicone sealant on all threaded connections.  
Threaded Connections:  
1.  
Prior to performing this maintenance on any threaded fitting, release the pressure  
on the piping/tubing system (i.e., turn off the pump, open a drain port, etc.)  
2.  
3.  
Remove the connecting tubing.  
Re-tape the pipe threads with 2 or 3 wraps of Teflon tape. (It is not necessary to  
remove the old Teflon tape, although it may be advisable to do so if it is in very  
bad shape).  
4.  
Re-insert the threaded fitting into the opening and tighten hand tight. Be careful  
to avoid cross threading. Tighten further using a wrench only if necessary, and  
only about one-half turn.  
5.  
6.  
Re-connect piping, pressurize, and inspect for leaks.  
If the leak is still present, de-pressurize system again and tighten the fitting  
another quarter-turn. Check again for leaks. (Over tightening can crack the  
female fitting).  
7.  
If the leak continues, it may be necessary to replace the male threaded fitting  
and/or the female threaded fitting.  
14693e - 31 Mar 03  
4- 6  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Tubing connections:  
14693e - 31 Mar 03  
4- 7  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
1.  
Prior to performing this maintenance on any tube fitting, release any pressure on  
the tubing system (i.e., turn off the pump, open a drain port, etc.).  
2.  
3.  
Remove the red locking clip (if applicable).  
Push the dark gray collar in towards the body of fitting (light gray). This releases  
the ‘gripper’ which holds the tubing in place.  
4.  
5.  
Hold the gray collar in while gently pulling the tube away from the fitting. It may  
be helpful to wiggle the tube side-to-side slightly while pulling.  
Remove the gripper collar by pulling straight out. Reach into the opening with a  
small, blunt (non-scratching) probe and remove the O-ring. (Inspect the collar to  
insure that all grippers are intact.)  
6.  
7.  
Rinse the O-ring with warm running water to remove any dirt, debris, or crusty  
concentrate and set aside to dry on a clean paper towel.  
Clean the inside of the fitting with a wet cotton swab or paper towel to remove  
any dirt or debris. Be sure not to leave any cotton fibers from the swab or towel  
inside the fitting.  
8.  
Lubricate the O-ring sparingly with silicone O-ring lubricant and place the O-ring  
back inside the fitting.  
9.  
Replace the gripper collar.  
10.  
With a sharp utility knife, cut off approximately ½” to 1” of the tube. Carefully  
cut the tube as square (perpendicular to the length) as possible. Inspect remaining  
tubing for marks/scratches and replace/repair as necessary.  
11.  
12.  
Place the tube in the fitting opening and push firmly until the tube seats fully in  
the socket (completely ‘bottomed out’).  
Pull back on the tubing to seat the grippers, and re-install the locking clip (if  
applicable) between the light gray body of the fitting and the dark gray collar (flat  
side towards the body).  
13.  
14.  
Re-pressurize the system (if applicable) and inspect for leaks.  
If the leak is not resolved, start over or simply replace the entire tube fitting or  
tube.  
14693e - 31 Mar 03  
4- 8  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.2.4 Product Recirculation Check Valve Test  
With Valve In RO Unit  
1.  
Measure TDS of Product water using a handheld meter. If RO display and TDS  
measurement do not match within a small amount, replace check valve. If values  
match, check valve is functioning correctly.  
2.  
Measure and record the Product water flow (using a flow meter or graduated  
cylinder and stopwatch). After determining flow rate, slowly kink the waste line  
being careful to not close off the flow completely. Measure Product water flow  
again and compare against previous results. If there is a noticeable increase in  
flow, the check valve needs replaced.  
3.  
If none of the above occurs, check valve is functioning correctly. For further  
verification, proceed with the following test.  
With Valve Removed From RO Unit  
1.  
Apply pressure in opposite direction of flow indicated on check valve. Check  
pressure reading (using a gauge) and for seepage. Remove pressure from valve  
and cycle it (push in direction of flow several times using care not to scratch or  
damage valve). Reapply pressure in opposite direction of flow and recheck  
pressure and for seepage. If pressure does not drop and no seepage is found,  
valve is functioning correctly.  
14693e - 31 Mar 03  
4- 9  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.3  
Millenium™ Cleaning Indications  
Foulants such as minerals, silt and organic substances may collect on the RO membrane  
surface and cause the RO membrane to foul and diminish in its function. Many of these  
foulants are prevented by proper selection of pretreatment equipment, while others are  
treatable only with more sophisticated designs. Regardless of the foulant, it is essential  
to clean the membrane promptly once cleaning is indicated. The longer a foulant is  
allowed to remain in contact with the RO membrane, the more difficult it will be to  
remove. In severe cases, repeated cleanings may be required.  
NOTE:  
Some foulants CANNOT be removed with cleaning and membrane performance  
may not be fully restored. Observe the appearance of the cleaning solution before  
discarding it. If the solution is heavily soiled, repeat the cleaning procedure with  
fresh solution of the same type (high pH or low pH).  
Routine cleaning of the RO membranes in conjunction with regular disinfection will keep  
bacteria counts at a minimum and the RO membrane performance at its best. Cleaning  
removes the silt, organic and mineral deposits that collect on the RO membrane surface  
which, if left untreated, encourage bacterial growth and will eventually impair the RO  
membranes' ability to perform. Disinfection kills most of the bacteria that may be present  
in the system, but it does not remove the deposits on the membrane.  
There are a variety of indications for cleaning membranes. The most common is a  
reduced product flow rate. Remember that a reduction in inlet water temperature can also  
result in a reduced flow rate. Refer to Technote 113 "Temperature Correction Factors" to  
determine the reduction in flow that can be attributed to reduced water temperature.  
Clean the membranes whenever the temperature compensated flow rate drops by 10%. A  
loss in the membrane performance (percent rejection) or increase in TDS of the product  
water may also be an indication to clean the membranes. A good preventative  
maintenance practice is to clean the membranes on a quarterly basis, and more often if  
needed.  
CAUTION:  
Membranes that have not been cleaned for a long time may actually incur damage  
during cleaning as perforations are exposed that were masked by accumulations of  
foulant. Regular maintenance cleaning is the best assurance of long membrane  
service.  
14693e - 31 Mar 03  
4- 10  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.3.1  
Pre-Cleaning Preparation  
There are two types of cleaners available, a low pH cleaner (Calcium/Iron removal) and a  
high pH cleaner (Silt/Organics removal). GE Osmonics provides these cleaners in liquid  
form. The low pH cleaner attacks mineral scale deposits such as calcium, magnesium  
and iron build-up, and should be used first. The high pH cleaner will remove  
accumulated silt and organic matter from the RO membrane.  
CAUTION:  
The different cleaners must not be mixed together. Introduce the low pH cleaner  
first, thoroughly rinse the system, and then use the high pH cleaner.  
WARNING:  
The solution is a mild irritant to eyes and skin. It is recommended that protective  
gloves and eye protection be worn when handling the cleaner. Neither respiratory  
protection nor special ventilation are required.  
Materials Required:  
Liqui-Clean 403™ – (Low pH liquid) - GE Osmonics Part No. 1228653 (1 Gal)  
Liqui-Clean 211™ – (High pH liquid) - GE Osmonics Part No. 1228791 (1 Gal)  
high and low pH cleaner  
2 Gallons pretreated water  
clean wide mouth container with volume of 2 ½ gallon or more  
pH test strips  
product rinse line (clear hose with male QD fitting on one end, open on the other)  
“DO NOT USE” warning labels (not provided by Osmonics)  
cleaning / disinfect log sheets  
most current log sheets for water quality comparison  
4.3.1.1 Pre-Cleaning Procedure  
Read and understand instructions on cleaner container before beginning procedure.  
1.  
2.  
3.  
Connect the water inlet line to the water supply, and the waste and product lines  
to drain. Turn the RO unit and water supply on.  
Establish a pre-cleaning % rejection, TDS and pH record on the cleaning disinfect  
log sheet.  
Prepare the cleaning solution by adding the amount of cleaner as specified on the  
labeling to two gallons of pretreated or RO product water.  
14693e - 31 Mar 03  
4- 11  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
CAUTION:  
Water to be used must be chlorine free or damage to the RO membranes may result.  
4.  
Turn the RO pump power and inlet water valve switches off.  
WARNING:  
Label machine with appropriate warning signs such as "Do Not Use / Contains  
Cleaner" to prevent use of the RO until it is properly rinsed.  
5.  
6.  
7.  
Connect the disinfect hose quick-disconnect fitting to the disinfect port on the  
RO. Place the open end in the cleaning solution.  
Connect the product rinse line to the product port and place the open end in the  
cleaning solution.  
Place the open end of waste line in the cleaning solution.  
4.3.1.2 Cleaner Recirculation and Dwell Time  
1.  
Turn the Operate/Flush valve to the Flush position. The Flush mode will allow the  
flow of cleaner only to the rejection side of the RO membrane to flush surface  
particulates away. Very little or no product water will be made in flush mode.  
2.  
Push and hold the disinfect switch to operate the RO pump for three minutes.  
CAUTION:  
Some pump noise (chattering) is normal in the flush position. Excessive pump  
noise (chattering) can be damaging to the pump. The RO is in “flush” from the full  
open (flush) position to just before the fully closed (operate) position. Adjust it by  
partly closing the flush valve to reduce chattering of the pump.  
CAUTION:  
DO NOT ALLOW PUMP TO RUN DRY. Damage will occur if pump is run dry.  
CAUTION:  
Check that there is waste flow within 10 seconds. If not, release the button (switch)  
and verify that the disinfect-line is connected properly.  
CAUTION:  
Do not allow air or foam to be drawn into the RO system, as this will damage the  
pump.  
NOTE:  
The cleaners will foam; this is normal. If using a closed container, provide a vent  
to the atmosphere to relieve pressure.  
3.  
4.  
5.  
Let the RO stand dormant with cleaning solution in it for thirty minutes.  
Remove the waste hose from the cleaning solution and place it in the drain.  
Using the disinfect switch, turn the RO on and flush approximately 1/2 gallon of  
the solution to the drain. This prevents fouling the cleaning solution with the first  
14693e - 31 Mar 03  
4- 12  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
output after the dwell time.  
6.  
7.  
8.  
Place the waste hose back in the cleaning solution container.  
Using the disinfect switch, turn the RO on and operate it for five minutes.  
Remove the product and waste hoses from the cleaning solution tank and place  
them in the drain.  
4.3.1.3 Post-Cleaning Rinse Procedure  
1.  
Using the disinfect switch, turn the RO on and run cleaning solution to the drain  
until the container is empty. This aids in emptying the container of cleaning  
solution.  
CAUTION:  
DO NOT ALLOW THE PUMP TO RUN DRY. As an alternative, the solution  
may be poured in the drain.  
2.  
3.  
4.  
Rinse and drain the cleaning solution tank and the hoses then let them air dry.  
Disconnect the disinfect hose from the RO.  
Ensure that the water supply to the RO unit is on. If it is partially closed, rotate  
the operate/flush valve to the full flush position.  
5.  
Place the inlet water valve switch in the on position. Let water flow to the drain  
for one minute.  
6.  
7.  
Rotate the operate/flush valve to the operate position.  
Set the RO pump power switch in the ON position. Rinse for 20 minutes with the  
Operate/Flush valve in the Operate position until the % rejection and TDS  
monitors’ displays normalize (e.g. 'usual' % rejection and 'usual' TDS reading).  
Refer to your operating log for 'usual' readings and compare to the pre-cleaning  
values.  
A. During this rinse period, rotate the Operate/Flush valve to the  
Flush position briefly and then back to Operate three times to  
build-up and release pump pressure. This aids in flushing the  
cleaner from the system.  
8.  
Test the product water for post-cleaning pH and compare it to the pre-cleaning pH  
level. The values should be the same or very close to the pre-cleaning pH. If it is  
not, continue rinsing in the Operate position until equalization of the value occurs.  
9.  
When all of the parameters (% Rejection, TDS, pH) have normalized, operate the  
RO for an additional 20 minutes in rinse mode.  
10.  
Repeat the cleaning procedure with the other pH cleaner. GE Osmonics  
recommends using a low pH cleaner first, followed by a high pH cleaner after  
rinsing thoroughly.  
14693e - 31 Mar 03  
4- 13  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
11.  
Turn the RO pump and inlet water valve switches OFF. Disconnect the product  
rinse line, and drain and store it.  
12.  
13.  
14.  
Reconnect the product line to its original connection.  
Remove the “Do Not Use/Contains Cleaner” warning label.  
The Millenium™ RO system is now ready for use.  
14693e - 31 Mar 03  
4- 14  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.3.2  
BIOSAN™ Pre-Cleaning Preparation  
BIOSAN™ is a liquid cleaner/descaler agent intended for use in removing mineral scale  
and organic biofilm and is a short-term storage solution for RO water systems.  
CAUTION:  
BIOSAN™ is not intended for use in Hemodialysis Machines.  
WARNING:  
The solution is a mild irritant to eyes and skin. It is recommended that gloves  
(latex/vinyl) and eye protection be worn when handling BIOSAN™. Neither  
respiratory protection nor special ventilation are required.  
Materials Required:  
BIOSAN™ cleaner  
product water  
clean wide mouth container with volume of 2 ½ gallon or more  
pH test strips  
product rinse line (clear hose with male QD fitting on one end, open on the other)  
“DO NOT USE” warning labels (not provided by Osmonics)  
cleaning / disinfect log sheets  
most current log sheets for water quality comparison  
4.3.2.1 BIOSAN™ Cleaning Procedure  
Read and understand all BIOSAN™ cleaning instructions before beginning  
procedure.  
1.  
2.  
3.  
4.  
Connect the water inlet line to the water supply, and the waste and product lines  
to drain. Turn the RO unit and water supply on.  
Establish the pre-cleaning % rejection, pH and TDS readings and record them on  
the Cleaning/Disinfect Log Sheet.  
Prepare the BIOSAN™ solution in accordance with the manufacturers  
instructions.  
Turn the RO pump power and inlet water valve switches off.  
WARNING:  
Label machine with appropriate warning signs such as "Do Not Use / Contains  
BIOSAN™" to prevent use of RO until properly rinsed.  
CAUTION:  
Water to be used must be chlorine free or damage to the RO membranes may result.  
14693e - 31 Mar 03  
4- 15  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
5.  
6.  
7.  
Connect the disinfect line quick-disconnect fitting to the disinfect port on the RO.  
Place the other (open) end of the line in the cleaning solution.  
Connect the product rinse line to the product QD port and place the other end of  
the line in the cleaning solution.  
Place the end of the waste line in the cleaning solution.  
4.3.2.2 BIOSAN™ Recirculation and Dwell Time  
1.  
2.  
Rotate the Operate/Flush valve to the Operate position. This will allow the  
BIOSAN™ to pass from the rejection side of the membrane through to the  
product side of the membrane for total system exposure to the cleaning solution  
once the RO is turned on in disinfect mode (step #2).  
Push and hold the disinfect switch to operate the RO pump for three minutes.  
CAUTION:  
Check that there is waste flow within 10 seconds. If not, press the disinfect switch  
to turn RO pump off and verify that the disinfect line is connected properly.  
CAUTION:  
DO NOT ALLOW PUMP TO RUN DRY. Damage will occur if pump is run dry.  
CAUTION:  
Do not allow air or foam to be drawn up into the RO unit, as this will damage the  
pump.  
NOTE:  
Biosan™ will foam; this is normal. If using a closed container, provide a vent to  
the atmosphere to relieve pressure.  
3.  
Turn the Operate/Flush valve to the Flush position and hold the disinfect switch  
for an additional 3 minutes. The Flush mode will allow the flow of cleaner only to  
the reject side of the RO membrane to flush particles away. Very little or no  
product water will be made.  
CAUTION:  
Some pump noise (chattering) is normal in the flush position. Excessive pump  
noise (chattering) can be damaging to the pump. The RO is in “flush” from the full  
open (flush) position to just before the fully closed (operate) position. Adjust by  
partly closing the flush valve to reduce the chattering of the pump.  
4.  
Allow the BIOSAN™ to dwell in the RO for a minimum of one hour.  
14693e - 31 Mar 03  
4- 16  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
NOTE:  
Thin-film RO membranes: The Millenium™ RO unit may be stored in BIOSAN™  
for up to two weeks.  
5.  
Remove the waste and product lines from the cleaning solution and place them in  
the drain.  
4.3.2.3 BIOSAN™ Post-Cleaning Rinse Procedure  
1.  
Using the disinfect switch, turn the RO on and run the cleaning solution to the  
drain until the container is empty.  
CAUTION:  
DO NOT ALLOW THE PUMP TO RUN DRY. As an alternative, the solution  
may be poured in the drain.  
2.  
3.  
Disconnect the disinfect line from the RO.  
Ensure that the water supply to the RO system is on. If it is partially closed, rotate  
the operate/flush valve to the full flush position.  
4.  
Push the inlet water valve switch to the ON position. Let water flow to the drain  
for one minute.  
5.  
6.  
Rotate the Operate/Flush valve to the operate position.  
Set the RO pump power switch in the ON position. Rinse for 20 minutes with the  
Operate/Flush valve in the Operate position until the % rejection and TDS  
monitor displays normalize (e.g. 'usual' % rejection and 'usual' TDS reading).  
Refer to your operating log for 'usual' readings and compare to the pre-cleaning  
values.  
A. During this rinse period, rotate the Operate/Flush valve to the Flush position  
briefly and then back to Operate three times to build-up and release pump  
pressure. This aids in flushing the cleaner solution from the system.  
7.  
Test the product water for post cleaning pH and compare it to the pre-cleaning pH  
level. The values should be the same or very close. If it is not, continue rinsing  
in the Operate position until equalization of the value occurs.  
8.  
9.  
When all of the parameters (% Rejection, TDS, pH) have normalized, operate the  
RO for an additional 20 minutes in rinse mode.  
Turn the RO pump and inlet water valve switches OFF. Disconnect the product  
rinse line, and drain and store it.  
10.  
11.  
12.  
Reconnect the product line to its original connections.  
Remove the 'Do Not Use/Contains BIOSAN™' warning label.  
The Millenium™ RO System is now ready for use.  
14693e - 31 Mar 03  
4- 17  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.4  
Millenium™ Disinfection Indications  
DETERMINATION OF MICROBIAL CULTURING AND DISINFECTION  
FREQUENCY IS THE RESPONSIBILITY OF THE OPERATING FACILITY.  
The following is recommended for critical applications where bacteria and endotoxins  
are a major concern, such as for Hemodialysis.  
Disinfection should occur when bacterial growth is above the limits set by the facility, or  
when over 200 CFU/ml (maximum AAMI standard) is detected. Routine disinfection at  
lower CFU counts will inhibit colonization of bacteria in the system and the distribution  
plumbing.  
Rapid re-growth of bacteria following disinfection, or excessively high colony counts  
(i.e. TNTC), may indicate a need for more aggressive intervention. Strategies such as  
increased frequency of disinfection, longer dwell times, or alternative chemical agents  
may be employed.  
CAUTION:  
Glutaraldehyde and Actril™ are not recommended disinfectants for the  
Millenium™.  
4.4.1  
Formaldehyde Disinfection Preparation  
CAUTION:  
New replacement thin film membranes must be rinsed to drain with normal  
pressures for a minimum of six (6) hours before first exposure to formaldehyde, or  
irreversible damage may occur. No mandatory pre-rinse of the membranes is  
required for future disinfections.  
WARNING:  
Always wear rubber gloves, apron and facemask. Use proper ventilation during  
disinfection and rinse-out. A face respirator with an organic cartridge to remove  
formaldehyde fumes may be worn. DO NOT BREATHE FUMES OR ALLOW  
FORMALDEHYDE TO COME IN CONTACT WITH SKIN OR EYES. Refer to  
MSDS supplied by manufacturer of the formaldehyde or formalin product.  
Materials Required:  
800 ml. Formaldehyde, 37%  
2 gallons pre treated or RO water  
clean wide mouth container with volume of 2 ½ gallon or more  
formaldehyde test kit  
product rinse line (clear hose with male QD fitting on one end, open on other)  
“DO NOT USE” warning labels or signs (not provided by Osmonics)  
cleaning / disinfect log sheets  
most current log sheets for water quality comparison  
proper protective body wear  
WARNING:  
14693e - 31 Mar 03  
4- 18  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Label the machine with appropriate warning signs such as "Do Not Use / Contains  
Formaldehyde" to prevent use of the RO until it is properly rinsed.  
4.4.1.1 Formaldehyde Pre-Disinfection Procedure  
1.  
2.  
3.  
Turn the Millenium™ RO off.  
Disconnect the RO system from the dialysis machine.  
Connect the RO product rinse line to the product port and place the open end in a  
drain.  
4.  
Connect the waste line to the RO and run the other end to a drain.  
NOTE:  
Follow your facility’s procedure for formaldehyde disposal. The drain line may  
have to be placed in a container. Verify that the operate/flush valve is in the  
OPERATE position.  
5.  
Prepare a 4% solution of formaldehyde (800 ml of 37% formaldehyde to 2 gallons  
of pre-treated water. The 2 gallons of solution is adequate for the Millenium™  
RO).  
CAUTION:  
Formaldehyde or Formalin is usually provided in a clinical setting in a 37%  
solution. Check the labeling on the packaging to make certain this is the case. If the  
solution is other than 37%, the dilution will require adjustment. The end solution  
should not exceed 4.2%. Consult the Medical Director or Director of Nursing to  
determine the correct solution strength. Osmonics will not be responsible for any  
damaged components due to incorrect solution strength.  
6.  
Attach the quick-disconnect end of the disinfect line to the disinfect port of the  
RO. Place the open end of the disinfect line in the disinfectant.  
4.4.1.2 Formaldehyde Disinfection and Dwell Time  
1.  
Push and hold the disinfect switch to operate the RO pump. Allow the RO to  
operate until the disinfectant has been drawn into the machine.  
CAUTION:  
Check that there is waste flow within 10 seconds. If not, push the switch once and  
verify that the disinfect line is submerged in the solution and the QD is connected  
correctly. Do not allow air to be drawn into the RO system, as this will damage the  
pump.  
2.  
3.  
Test for the presence of formaldehyde in the product line (the RO must be  
running).  
Release the disinfect switch.  
14693e - 31 Mar 03  
4- 19  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.  
Momentarily turn the inlet water valve switch on (approximately 5 seconds), then  
turn it off again. This allows the water to rinse the disinfectant from the disinfect  
line and quick-disconnect connection.  
WARNING:  
Failure to perform this step creates a potentially dangerous situation  
(FORMALDEHYDE SPILLAGE) and prolongs the rinsing out of the disinfectant.  
5.  
6.  
7.  
8.  
Disconnect the disinfect line from the quick-disconnect port.  
Ensure that the FORMALDEHYDE warning label is clearly visible.  
Remove the product rinse line and waste line from disinfect tank and RO.  
Rinse the disinfect line and the disinfect tank with water to remove the  
formaldehyde, and store it appropriately.  
9.  
Allow the disinfectant solution in the RO to remain in contact for a minimum of  
two hours, or, preferably, overnight.  
NOTE:  
The RO may be stored for up to one year at this time. If this is the case, a 0.5%  
solution may be used instead of 4%.  
4.4.1.3 Formaldehyde Post-Disinfection Rinse Procedure  
1.  
2.  
3.  
4.  
5.  
Connect the product rinse line, input water and waste lines to the RO.  
Ensure that the product rinse and waste lines are in the drain.  
Turn the operate/flush valve to the flush position.  
Turn on the inlet water valve and allow the system to rinse for 2 minutes.  
Turn on the RO power switch and run the RO for 10 minutes.  
CAUTION:  
Some pump noise (chattering) is normal in the flush position. Excessive pump  
noise (chattering) can be damaging to the pump. The RO is in “flush” from the full  
open (flush) position to just before the fully closed (operate) position. Adjust it by  
partly closing the flush valve to reduce the chattering of the pump.  
6.  
7.  
Rotate the operate/flush valve to the operate position.  
Turn RO pump power switch off for 1 minute. Turn the RO pump switch on for  
2-3 minutes. This will aid in the rinsing out any trapped formaldehyde in the  
Millenium™ RO.  
8.  
9.  
Continue to rinse for an additional 5 minutes.  
Check the product line for residual disinfectant with an appropriate test kit. If the  
test is positive, continue to rinse until a negative test is obtained.  
10.  
Disconnect the product rinse line and store it.  
14693e - 31 Mar 03  
4- 20  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
11.  
Remove the warning signs from the RO only after a negative test confirms the  
absence of formaldehyde.  
12.  
The Millenium™ RO System is now ready for use.  
NOTE:  
Success of the disinfection procedure may be ascertained by taking a water culture  
after completion. See the Product Water Culture Procedure in this chapter.  
14693e - 31 Mar 03  
4- 21  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.4.2  
Peracetic Acid/Peroxide-Based Compounds (Renalin®) Disinfection  
Procedure  
Renalin®, Minncare™, Peracidin™ and others are trade names for compounds based  
on Peracetic Acid and Hydrogen Peroxide compounds, and are common in dialysis  
facilities. The products referred to in this procedure typically have 20% hydrogen  
peroxide and 4% peroxacetic acid before dilution. Other products are available with  
these same chemicals in different ratios; these are not acceptable for use in the  
Millenium™ RO. Actril™ and other compounds with different combinations of  
these chemicals must not be used on polyamide (thin-film) membranes. The name,  
“Renalin®” will be used in this manual to refer to acceptable products.  
CAUTION:  
New replacement thin film membranes must be rinsed to drain with normal  
pressures for a minimum of three (3) hours before first exposure to Renalin. No  
mandatory pre-rinse of the membranes is required for future disinfections.  
WARNING:  
Follow all Material Safety Data Sheet (MSDS) precautions and OSHA Standards.  
CAUTION:  
‘RENALIN®’ is corrosive to most metals other than stainless steel. The  
Millenium™ uses only compatible materials. ‘RENALIN®’ type products contain  
oxidizing agents and are potentially damaging to thin film (polyamide) membranes.  
It is imperative that all the precautions and instructions are followed correctly.  
Failure to comply with these instructions, which incorporate instructions from the  
membrane manufacturer, will jeopardize the system performance and the warranty.  
Materials Required:  
correct amount of disinfectant, per manufacturers’ instructions  
2 gallons RO water  
clean wide mouth container with volume of 2 ½ gallon or more  
product rinse line (clear hose with male QD fitting on one end, open on the other)  
“DO NOT USE” warning labels or signs (not provided by Osmonics)  
cleaning / disinfect log sheets  
most current log sheets for water quality comparison  
14693e - 31 Mar 03  
4- 22  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.4.2.1 Renalin® Pre-Disinfection Preparation  
CAUTION:  
When ‘RENALIN®’ is to be used as a disinfectant, several preparatory steps are  
required by GE Osmonics as well as the membrane manufacturer. Failure to  
comply with these instructions, which incorporate instructions from the membrane  
manufacturer, will jeopardize the system performance and warranty.  
a.  
b.  
The system must be free of all incompatible materials.  
Clean the membranes with an acid such as 0.1% HCl or 0.4% phosphoric  
acids. Calcium and Iron Cleaner may be used as an alternate. It is essential  
that all iron and other transition metals be removed from the membrane  
surface.  
CAUTION:  
The RO system must be disconnected from the tempered water source (if present).  
The solution temperature must be monitored closely. At temperatures above 25°  
the ‘RENALIN®’ solution becomes very aggressive towards the RO membranes  
and will damage them in a short period of time. The disinfection procedure will  
raise the temperature of the solution. The operator must be prepared to add ice to  
the solution or discontinue the procedure when the water temperature approaches  
77ºF /25°C.  
1.  
Read and understand the ‘RENALIN®’ disinfection procedure. Pay particular  
attention to steps a and b in CAUTION statement above before initiating the  
procedure.  
2.  
3.  
Ensure the cleaning/disinfectant container is clean. Add two gallons (7.6 liters) of  
RO permeate (product water) to the container.  
Add 2.6 ounces (80 milliliters) of disinfectant concentrate to the container and  
mix to prepare a 1:100 (1%) dilution.  
WARNING:  
Label the machine with appropriate warning signs such as "Do Not Use/ Contains  
Disinfectant” to prevent use of the RO until properly rinsed.  
4.  
Disconnect the dialysis machine from the RO product port and install the product  
rinse line. Place the open end in the disinfectant solution container and ensure the  
product valve is open.  
5.  
6.  
Attach the quick-disconnect end of the disinfect line to the disinfect port of the  
RO. Put the open end of the disinfect line in the disinfectant.  
Disconnect the waste line from the drain connection and place the open end in the  
disinfectant solution container.  
14693e - 31 Mar 03  
4- 23  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.4.2.2 Renalin® Disinfection and Dwell Time  
1.  
2.  
Verify that the operate/flush valve is in the OPERATE position.  
Use the disinfect switch to circulate the solution through the system for five  
minutes. After three minutes of circulation, test the product line for the presence  
of disinfectant. Continue circulating until a positive test is achieved.  
CAUTION:  
Do not allow the solution temperature to climb above 25°C during this step. Failure  
to control temperature will result in rapid deterioration of the membrane, resulting  
in poor rejection.  
3.  
4.  
5.  
Place the product and waste lines into a drain. Protect the product line from  
contamination.  
Using the disinfect switch, turn the RO on and run solution to the drain until the  
container empties. Alternately, the container may be carefully poured in the drain.  
Turn off the RO; do not operate the RO dry.  
CAUTION:  
Do not allow air to be drawn up into the RO unit as this will damage the pump.  
6.  
Momentarily turn the inlet water valve switch on (approximately 5 seconds.) This  
allows the water to rinse the disinfectant from the disinfect line and quick-  
disconnect connection.  
WARNING:  
Failure to perform this step creates a potentially dangerous situation  
(DISINFECTANT SPILLAGE) and prolongs the rinsing out of the disinfectant.  
7.  
8.  
Remove the disinfect line from the disinfect port. Rinse, drain and store it.  
Allow the RO to soak in the disinfecting solution for 36 minutes -12 hrs. Thirty-  
six minutes contact time is expected to result in 90% bacteria kill; 12 hours is  
expected to kill 99% of the bacteria. The minimum amount of soak to achieve  
disinfection is better for the membranes.  
4.4.2.3 Renalin® Post-Disinfection Rinse Procedure  
1.  
2.  
3.  
4.  
5.  
Ensure that the product rinse line is connected to the RO.  
Ensure that the product rinse and waste lines are connected to the drain.  
Rotate the operate/flush valve to the flush position.  
Turn on the inlet water valve switch and allow the system to rinse for 2 minutes.  
Turn the RO pump power switch on and run the RO for 10 minutes.  
CAUTION:  
14693e - 31 Mar 03  
4- 24  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
The operate/flush valve may need to be rotated partially closed, (towards the  
operate position) if the pump makes excessive noise.  
6.  
7.  
Rotate the operate/flush valve to the operate position.  
Turn the RO pump power switch off for 1 minute. Turn the RO pump switch on  
for 2-3 minutes. This will aid in the rinsing out of any potentially trapped  
disinfectant in the Millenium™ RO.  
8.  
9.  
Continue to rinse for an additional 5 minutes.  
Check the product line for residual disinfectant with an appropriate test kit. If the  
test is positive, continue to rinse until a negative result is achieved.  
10.  
Disconnect the product rinse line and store it.  
11.  
Remove the 'Do Not Use/Contains Disinfectant’ warning label. The  
Millenium™ RO System is now ready for use.  
14693e - 31 Mar 03  
4- 25  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.5  
Product Water Culture  
FREQUENCY: Once a month or more often, depending upon culture results, and unit  
protocol. We recommend before and after disinfection.  
Materials Required:  
Bucket or similar container  
Sterile specimen container, such as Millipore (red) or 3M Commercial Standard  
Plate count sampler, or a container your lab may require  
RATIONALE: This procedure will help ensure that the microbial count of the product  
(RO) water falls within the AAMI Hemodialysis Water Standard. The AAMI standard  
maximum allowable concentration for bacteria in product water is 200 colony-forming  
units. A bacterial count above the maximum level may be harmful to the hemodialysis  
patient.  
OTHER: Consult your lab to find out what type of sterile container to use and how  
much product (RO) water is required to perform the test. Make certain that your  
laboratory is informed of the correct procedure for performing a product water culture. A  
"spread plate method" NOT a "calibrated loop" technique should be used, with tryptic  
soy agar or equivalent agar, NOT blood or chocolate agar. If tested incorrectly, the  
results may be inaccurate. The sample should be assayed within 30 minutes of obtaining  
it, or refrigerated (5°C) immediately and assayed within a 24 hour period. Obtain the  
samples wearing long sleeves and a mask to prevent contamination of samples, use a  
“mid-stream” “clean catch” type procedure.  
14693e - 31 Mar 03  
4- 26  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.5.1 Product Water Culture Procedure  
This procedure assumes the RO is connected to power and water, and is in operating  
condition.  
1.  
2.  
3.  
4.  
5.  
Connect the product rinse line to the product port. Place the other end in a drain.  
Connect the waste line to the waste port. Place the other end in a drain.  
Turn on the RO unit. Allow the RO unit to run for 10-15 minutes.  
Remove the product rinse line from the drain.  
Using aseptic technique, hold the product line over a container, and allow  
approximately 100 cc of product (RO) water to flow to the drain (or allow product  
water to flow for 30 seconds).  
6.  
7.  
Maintaining aseptic technique, hold the product line over the sterile container  
with the product water flowing.  
Fill the container with RO product water. (If using commercial samplers, follow  
the manufacturer's instructions.)  
8.  
9.  
Aseptically cap the specimen container immediately.  
Label the specimen appropriately with:  
Test to be performed - "culture / colony count"  
Sample source - product water/RO, machine serial number and room location  
Time and date sample obtained  
Person who obtained specimen  
Any other pertinent information or procedures your facility or lab requires  
10.  
The samples should be assayed within 30 minutes of obtaining or refrigerated at  
5° C immediately and assayed within a 24-hour period.  
14693e - 31 Mar 03  
4- 27  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.6  
System Storage  
This storage procedure has been established to store Millenium™ RO machines. It is not  
recommended that a unit be dormant longer than 24 hours without disinfectant. ROs that  
have been left unused for more than 24 hours risk the possibility of colonizing bacteria  
and exceeding acceptable limits. ROs to be stored should first be disinfected with one of  
the procedures in this chapter. The RO should be stored in a location where it cannot  
freeze or become overly hot.  
CAUTION:  
‘RENALIN®’ products (Renalin®, Minncare™, Peracidin™. etc.) are not  
acceptable disinfectants for storage of systems. Exposure of more than 12 hours to  
RENALIN® will result in irreversible damage to the membranes.  
4.6.1 Formaldehyde Storage Procedure  
WARNING:  
Always wear rubber gloves, an apron and a facemask. If the room has ventilation,  
use it. If not, wear a face respirator with an organic cartridge to remove  
formaldehyde fumes. DO NOT BREATHE FUMES OR ALLOW  
FORMALDEHYDE TO COME IN CONTACT WITH SKIN OR EYES Refer to  
the MSDS supplied by the manufacturer of the formaldehyde/formalin product.  
1.  
If formaldehyde is chosen to disinfect the RO, follow the instructions for  
formaldehyde disinfection.  
2.  
3.  
Do not rinse out after the disinfectant has been circulated in the machine.  
Disconnect the disinfect line, the feed water line, the waste line, power and the  
product line.  
4.  
Coil the hoses and power cord, and cover the RO with a large plastic bag to  
protect the RO from dust.  
NOTE:  
All ROs can be safely stored in 0.5 to 4 % formaldehyde for up to one year.  
WARNING:  
After storage, the RO unit must be completely rinsed of formaldehyde, and a  
bacterial culture performed and interpreted before use. Perform AAMI analysis if it  
has been one year or more since last done.  
14693e - 31 Mar 03  
4- 28  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
4.6.2 BIOSAN™ Storage Procedure  
WARNING:  
It is recommended that gloves (latex/vinyl and eye protection be worn when  
handling BIOSAN™.  
1.  
2.  
3.  
4.  
5.  
Disinfect and rinse the RO with an acceptable disinfectant.  
Follow the BIOSAN™ procedure in this chapter.  
Do not rinse the RO out after BIOSAN™ has been circulated in the RO.  
Disconnect the disinfect line, waste line, feed water line, power and product line.  
Coil the hoses and power cord, and cover the RO with a large plastic bag to  
protect the RO from dust.  
CAUTION:  
The Millenium™ RO may be stored for up to two weeks in BIOSAN™. After two  
weeks, if longer storage is needed, the Millenium™ RO may be rinsed and re-  
packed with BIOSAN™.  
WARNING:  
After storage, the RO unit must be completely rinsed of BIOSAN™, and a bacterial  
culture performed and interpreted before use. Perform AAMI analysis if it has been  
one year or more since last done.  
14693e - 31 Mar 03  
4- 29  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
NOTES:  
14693e - 31 Mar 03  
4- 30  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO System  
CHAPTER FIVE:  
TROUBLESHOOTING  
14693e - 31 Mar 03  
5- 1  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
5- 2  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Table of Contents  
SECTION  
5.1  
TITLE  
PAGE  
5-5  
RO Fails to Run  
5.2  
RO Runs Intermittently  
Motor Runs Hot  
5-5  
5.3  
5-6  
5.4  
Water Produced While RO is Off  
Low Pump Pressure  
5-6  
5.5  
5-7  
5.6  
Pressure Gauge Vibrates Erratically  
Excess Product Pressure (too high)  
Product Flow Too Low  
Product Flow Too High  
Leaks  
5-7  
5.7  
5-7  
5.8  
5-8  
5.9  
5-8  
5.10  
5.11  
5.12  
5.13  
5.14  
5.15  
5-9  
Low Percent Rejection  
TDS Value High  
5-9  
5-10  
5-11  
5-11  
5-11  
RO Fails to Draw Disinfectant  
RO Pump Leaks  
TDS Value fluctuates  
14693e - 31 Mar 03  
5- 3  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
5- 4  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
5.1  
RO Fails to Run  
1.  
Power off.  
a. Check position of inlet and pump switches.  
b. Check power cord (plugged in).  
c. Check outlet and circuit breaker/fuse panel for power with voltmeter.  
d. Check fuses in RO.  
2.  
3.  
Improper line voltage.  
a. Verify that line and equipment voltages are compatible.  
Loose electrical connections.  
a. Check all wiring connections to verify solid connection.  
b. Verify that plug is securely connected to outlet.  
4.  
5.  
6.  
Inlet water valve switch not activated.  
a. Check position of switch.  
b. Check switch with voltmeter.  
c. Replace switch.  
Power switch inoperative.  
a. Assure inlet water valve is open.  
b. Check switch with voltmeter.  
c. Replace switch.  
Pump motor inoperative.  
a. Verify that voltage supply is correct and motor is wired for correct voltage.  
Correct as necessary.  
b. Motor thermal overload switch tripped (most motors have automatic internal  
reset). Motor will restart when it cools approximately 10-15 minutes. Correct  
reason for the overload.  
c. Remove pump head and try restarting motor.  
d. Replace motor.  
7.  
Pump inoperative.  
a. Inspect pump and motor coupling for wear.  
b. Clean any debris or liquid from connection area.  
c. Replace pump.  
d. See 5.5.  
5.2  
RO Runs Intermittently  
1.  
Loose electrical connection.  
a. Verify that all wiring connections are secure.  
b. Verify that plug is securely connected to outlet.  
14693e - 31 Mar 03  
5- 5  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
2.  
3.  
Pump pressure too high.  
a. Adjust pressure at pump head.  
b. Check waste flow-path for obstruction.  
c. See 5.1.6  
d. See 5.1.7  
Thermal overload trips  
a. Verify line voltage and motor amperage draw.  
b. Disassemble pump/motor at clamp and inspect coupler for wear.  
c. Pump pressure too high (see 5.2.2).  
5.3  
Motor Runs Hot  
1.  
Pump/motor binding.  
a. Remove pump head and try restarting motor.  
b. Inspect pump and motor coupling for wear.  
c. Clean any debris or liquid from connection area.  
d. See 5.5  
e. Replace pump.  
2.  
3.  
Fan obstructed or inoperative.  
a. Verify no debris is obstructing fan vent.  
b. Verify fan operation.  
Excessive current draw.  
a. Verify that flow rates and pressures are within RO specifications.  
b. Check motor/pump coupling for alignment.  
c. Confirm amperage draw with meter.  
d. Motor insulation shorted, replace motor.  
4.  
Voltage too low.  
a. Check voltage against serial label and motor label, correct condition.  
5.4  
Water Produced While RO is Off  
1.  
Faulty inlet water solenoid valve.  
a. Cycle valve off and on with inlet water valve switch to clear obstruction.  
b. Remove valve and clear obstruction.  
c. Replace valve.  
2.  
Fault inlet water valve switch.  
a. Verify switch is in OFF position.  
b. Check switch with ohm meter, replace switch.  
14693e - 31 Mar 03  
5- 6  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
5.5  
Low Pump Pressure  
1.  
No pressure develops above level of feed pressure.  
a. Verify motor operation and water line connections.  
b. Verify flush valve is in the operate position.  
c. Verify (lower than normal) flow from product, if high see membrane problem.  
d. Verify (lower than normal) flow from waste, if high check flush valve operation.  
e. Pump damaged or worn out, replace pump.  
2.  
Pressure develops but does not achieve normal operating pressure.  
A. Waste or product flow too high.  
a. Check water temperature and flow rate.  
b. Check quality of product water.  
B. Faulty inlet pressure regulator.  
a. Replace regulator.  
C. Pump adjusted improperly  
a. Adjust pump at pump head and regulator.  
D. Pump worn out or damaged.  
a. Replace pump.  
5.6  
Pressure Gauge Vibrates Erratically  
1.  
Air in gauge tubing.  
a. Loosen tube fitting at gauge to bleed air.  
2.  
3.  
Obstruction in pressure gauge orifice.  
a. Remove obstruction.  
Product return check valve operation (Product Gauge Only)  
a. Check and remove obstruction in return check valve.  
b. Replace check valve.  
4.  
5.  
Faulty gauge.  
a. Replace gauge.  
Inlet regulator faulty.  
a. Replace regulator.  
5.7  
Excess Product Pressure (too high)  
1.  
Inlet pressure regulator improperly adjusted.  
a. Decrease pressure regulator to adjust maximum product setting.  
14693e - 31 Mar 03  
5- 7  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
2.  
Product flow rate too high.  
a. Decrease pump operating pressure if above specifications.  
b. Feed water temperature too high. Reduce temperature or discontinue use of RO  
to prevent damage to membranes.  
c. Damage or failure of membrane has occurred, replace membranes.  
d. Faulty product return check valve.  
5.8  
Product Flow Too Low  
Low water temperatures, fouled membrane elements, or inadequate pump pressure can  
cause low product flow. A partial blockage of the feed stream (dirty pre-filter) can also  
result in low product water flow by causing the pump to draw a strong vacuum.  
Similarly, an obstruction (defective check valve) in the product flow path may cause  
reduced flow.  
1.  
Pump pressure too low.  
a. See 5.5  
2.  
Low feed water temperature.  
a. Verify product production with Tech Note 121 Temperature Correction.  
b. Install or adjust temperature blending valve.  
3.  
Membrane fouled or scaled.  
a. Review trend analysis logs for flow.  
b. Clean membrane with both high and low pH cleaners.  
c. If flow is not recoverable, replace membrane.  
4.  
5.  
Center tube of ROMA collapsed from operating at high temperatures.  
a. Correct temperature problem.  
b. Replace membrane.  
Check softener pretreatment for brackish (high salt) water.  
a. Check inlet water for high conductivity.  
b. Check water softener for proper function.  
c. Regenerate softener.  
d. Check supply water for hardness and conductivity.  
6.  
Product recirculation check valve failed.  
a. Refer to Maintenance Procedures for Check Valve Test.  
14693e - 31 Mar 03  
5- 8  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
5.9  
Product Flow Too High  
1.  
Rejection Satisfactory  
A. Pump pressure too high.  
a. Adjust pressure regulator setting to specifications.  
B. Feedwater temperature too high.  
a. Adjust temperature blending valve.  
b. Correct plumbing error that allows the cold water to be heated.  
C. Quick disconnect tubing connections not in correct port.  
a. Verify QD connections.  
D. Product recirculation check valve failed.  
a. Refer to Maintenance Procedures for Check Valve Test.  
2.  
Rejection Unsatisfactory  
A. Waste flow too low.  
a. Check waste flow rate.  
b. Check for blockage or kink in waste line.  
c. Quick disconnect not connected completely.  
B. High pH and chlorines in feed water.  
a. Check feedwater for higher than 8.5 and the presence of chloramines.  
b. Check pH reduction equipment if applicable.  
c. Contact GE Osmonics.  
C. ROMA internal seals damaged, dirty, or dislodged.  
a. Disassemble ROMA, clean and inspect “O” rings and seals.  
D. Membrane hydrolyzed (exposed to oxidant).  
a. Verify condition of pretreatment equipment.  
E. Membrane damaged (or ruptured).  
a. Check for high product pressure.  
b. Check for product line being back pressured.  
5.10 Leaks  
1.  
General information: Any leak from any part of the system should be corrected as  
soon as possible. Left unattended, leaks can cause secondary damage  
necessitating expensive repairs.  
2.  
Refer to the Maintenance Procedures for Leak Repair.  
14693e - 31 Mar 03  
5- 9  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
5.11 Low Percent Rejection  
The overall membrane performance can be affected by a number of factors. Poor  
rejection performance, as well as high TDS, can result from fouled membranes, low  
pump pressure, inadequate waste flow, or feed water problems. An apparent  
performance problem could be caused by a defective sensor or circuit board, and not be a  
membrane issue at all. The recommended first step for any ‘membrane performance’  
question is to confirm the rejection percentage and product TDS with a hand-held  
RO/TDS meter. If the product water quality is poor, investigate the membrane.  
However, if the water quality tests ‘normal’, investigate the monitoring circuits.  
1.  
Waste flow too low.  
a. Disinfectant or cleaner not rinsed from RO.  
b. Check waste flow control.  
c. Check for obstruction or kink in waste line.  
2.  
Pump pressure too low.  
a. Adjust pressure regulator.  
b. Adjust pump settings.  
3.  
4.  
Product flow too high.  
a. See 5.9  
Membrane scaled or fouled.  
a. Verify condition of pretreatment equipment.  
b. See membrane cleaning procedure in this manual.  
5.  
Erroneous meter readout.  
a. Verify water quality with independent meter and check conductivity probe  
connection.  
b. Check all quick-disconnect connections.  
c. Replace inlet water sensor probe.  
d. Replace product water sensor probe (percent rejection).  
e. Replace monitor circuit board.  
6.  
7.  
Membrane failure.  
a. Replace membrane.  
Change in feed water TDS.  
a. Verify feed and product water quality with independent meter.  
b. Softener problem, rinsing out high levels of salt.  
14693e - 31 Mar 03  
5- 10  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
5.12 TDS Value High  
The overall membrane performance can be affected by a number of factors. Poor  
rejection performance, as well as high TDS, can result from fouled membranes, low  
pump pressure, inadequate waste flow, or feed water problems. An apparent  
performance problem could be caused by a defective check valve, sensor or circuit board,  
and not be a membrane issue at all. The recommended first step for any ‘membrane  
performance’ question is to confirm the rejection percentage and product TDS with a  
hand-held RO/TDS meter. If the product water quality is poor, investigate the  
membrane. However, if the water quality tests ‘normal’, investigate the monitoring  
circuits.  
1. Waste flow too low.  
a. Measure and compare to original data. Correct as necessary.  
2. Waste pressure too low.  
a. Adjust pump settings.  
3. Product flow too high.  
a. See 5.9  
4. Membrane scaled or fouled.  
a. Verify condition of pretreatment equipment.  
b. See membrane cleaning procedure in this manual.  
5. Erroneous meter readout.  
a. Verify water quality with independent meter.  
b. See 5.12.8  
c. TDS water probe wire disconnected or loose.  
d. Replace TDS water sensor probe.  
e. Replace thermistor.  
f. Replace monitor circuit board.  
6. Membrane failure.  
a. Replace membrane.  
7. Change in feed water TDS.  
a. Verify feed and product water quality with independent meter.  
b. Communicate with municipal water plant.  
c. Softener problem, rinsing out high levels of salt.  
8.  
Product recirculation check valve failed.  
a. Refer to Maintenance Procedures for Check Valve Test.  
14693e - 31 Mar 03  
5- 11  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
5.13 RO Fails to Draw Disinfectant  
1.  
Pre-filter bowl leaking on vacuum.  
a. Tighten bowl  
b. Inspect and clean pre-filter bowl seal / replace seal.  
c. Replace filter bowl  
2.  
3.  
Disinfect line kinked or plugged.  
a. Clear disinfect line.  
Air leak on disinfect draw line.  
a. Check that quick disconnect fitting is connected securely.  
b. Immerse tube in solution.  
c. Check tube for cracks or holes.  
d. Replace quick disconnect fitting.  
4.  
Solenoid failure in open position.  
a. See 5.4  
5.14 RO Pump Leaks  
1.  
2.  
Tubing connection leaking.  
a. Tighten or reconnect tubing fitting.  
Mechanical seal failure (leaks out vent hole).  
a. Replace pump.  
5.15 TDS Value Fluctuates  
1.  
Product recirculation check valve failed.  
a. Refer to Maintenance Procedures for Check Valve Test.  
14693e - 31 Mar 03  
5- 12  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
NOTES:  
14693e - 31 Mar 03  
5- 13  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
5- 14  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO System  
CHAPTER SIX:  
DRAWINGS  
14693e - 31 Mar 03  
6 - 1  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
6 - 2  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO  
(Component Layout)  
14693e - 31 Mar 03  
6 - 3  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO  
(Disinfect View)  
14693e - 31 Mar 03  
6 - 4  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO  
(Flow Schematic)  
14693e - 31 Mar 03  
6 - 5  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO  
(Electrical Schematic, High Voltage, 50/60Hz)  
14693e - 31 Mar 03  
6 - 6  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO  
(Electrical Schematic, Low Voltage, 50/60Hz)  
14693e - 31 Mar 03  
6 - 7  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
12312  
21080  
20204  
20205  
20203  
30005  
12310  
ELECTRICAL CONTROL PANEL  
30010  
1236062 (60Hz)  
1234687 (50Hz)  
12315 (60HZ)  
12315−1 (50HZ)  
12311  
ELECTRICAL CONTROL BOX  
Millenium RO  
(Electrical Layout)  
14693e - 31 Mar 03  
6 - 8  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
40569  
1234678  
Millenium RO  
(Input Manifold 50/60Hz)  
14693e - 31 Mar 03  
6 - 9  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
14765  
40218  
12131  
40569  
40564  
40602  
20178  
Millenium RO  
(Main Manifold)  
14693e - 31 Mar 03  
6 - 10  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
14738  
40976  
40976  
40976  
40976  
40976  
40218  
40570  
40218  
Millenium RO  
(Waste Manifold)  
14693e - 31 Mar 03  
6 - 11  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
OS1151581  
DETAIL ’A’  
(NOT TO SCALE)  
Millenium RO  
(ROMA Assembly)  
14693e - 31 Mar 03  
6 - 12  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
50112  
14493  
41566  
41566  
42014  
TO HOLE  
IN RO  
CABINET  
Millenium RO  
(Flush Valve Assembly)  
14693e - 31 Mar 03  
6 - 13  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
TOOLS REQUIRED:  
1. FLAT BLADE SCREWDRIVER.  
2. 3/4" WRENCH OR EQUIVALENT.  
3. TOWEL.  
PROCEDURE:  
1. TURN OFF POWER AND WATER TO UNIT.  
2. OPEN FRONT COVER AND EXPOSE PUMPHEAD.  
3. REMOVE ACORN NUT. NOTE: DO NOT LOSE WASHER.  
4. USE SCREWDRIVER AND TURN SCREW AS SHOWN. CLOCKWISE TO INCREASE PRESSURE, COUNTER  
CLOCKWISE TO DECREASE PRESSURE. EACH 90° ROTATION WILL CHANGE PRESSURE ABOUT 25 PSI.  
5. REPLACE WASHER (IF IT CAME OFF) AND ACORN NUT.  
6. TURN ON WATER AND POWER AND THEN CHECK PRESSURE READING. REPEAT AS NECESSARY.  
PRESSURE ADJ. SCREW  
DECREASE  
INCREASE  
WASHER  
ACORN NUT  
PUMP  
Millenium RO  
(Pump Pressure Adjustment Procedure)  
14693e - 31 Mar 03  
6 - 14  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
“Push-In” Fitting Instructions  
14693e - 31 Mar 03  
6 - 15  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
6 - 16  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
NOTES:  
14693e - 31 Mar 03  
6 - 17  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
6 - 18  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO System  
CHAPTER SEVEN:  
SPARE PARTS LIST  
14693e - 31 Mar 03  
7 - 1  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
This page intentionally left blank.  
14693e - 31 Mar 03  
7 - 2  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
Millenium RO System  
SPARE PARTS LIST  
NOTE: Part Numbers are subject to change and should be verified when ordering  
spare parts.  
CATALOG NUMBER  
DESCRIPTION  
PARTS  
42047  
Check Valve, Feed Line  
42047  
Check Valve, Product Line  
1228224  
1236062  
1234687  
30010  
Check Valve, Product Return  
Fuse, 1/4 AMP, Slo Blow  
Fuse, 1/8 AMP, Slo Blow  
Fuse Holder, Panel mount  
1234685  
20290  
Fuse, 8 AMP, Slo Blow  
Fuse, 5 AMP, Slo Blow  
1234348  
1234347  
1234349  
1234350  
30382  
Fuse drawer, 1 Pole  
Fuse drawer, 2 Pole  
Power cord retainer  
Inlet power module  
Power Cord (50Hz RO only, use w/ P/n 1236173)  
Plug End (50Hz RO only, use w/ P/n 30382)  
Gauge, Product Pressure  
1236173  
42048  
42074  
Gauge, Pump Pressure  
40179  
OS1228070  
30014  
35328  
35329  
Membrane, Tape Wrapped, for 500 GPD RO (2 required)  
Membrane, Tape Wrapped, for 750 GPD RO (2 required)  
Motor, 1/3 HP, 110/230 V, 50/60 Hz  
Fan and Motor, Cooling, 220VAC, 50Hz  
Fan and Motor, Cooling, 110VAC, 60Hz  
Power cord for Cooling Fan  
35330  
12315  
12315-1  
12310  
12311  
12312  
Power Supply PCB, 110/120 V, 60 Hz  
Power Supply PCB, 220/240 V, 50 Hz  
Printed Circuit Board, % Rejection  
Printed Circuit Board, Conductivity  
Printed Circuit Board, Display  
Disinfect Switch  
20204  
20205  
Lens for Disinfect Switch  
40165  
50041  
Pump, 100 GPH, SS, W/Regulator  
Pump-to-Motor Coupling  
1234489  
12131  
Regulator, Product Pressure Adjustment  
Sensor Probe, Total Dissolved Solids, Feed, Product  
14693e - 31 Mar 03  
7 - 3  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
CATALOG NUMBER  
DESCRIPTION  
PARTS (cont.)  
20203  
20203  
Switch, Inlet Water Valve  
Switch, RO Pump Power  
40574  
40739  
42014  
Valve, Feed Water Solenoid, 110 V, 60 Hz  
Valve, Feed Water Solenoid, 220 V, 50Hz  
Valve, Operate/Flush  
42014  
Valve, Product  
50003  
30335  
Clamp, Pump to Motor  
Selector Knob  
41611  
Selector Knob Shim  
OS1151419  
OS1151580  
OS1151581  
OS1160368  
41555  
Half Clamp for ROMA housing  
O-ring, Small  
O-ring, Large  
Endcap, Side port  
Hose Clamp, 5/8”  
1231491  
Terminal Butt Splice, no crimp  
FITTINGS  
40513  
40568  
Disinfect Port, Quick-disconnect (Female)  
Elbow, speed-fit, 3/8 x 3/8 mnpt, swivel  
Elbow, speed-fit, 3/8 x 3/8 stem  
40564  
41566  
Elbow, speed-fit, 1/4 x 3/8 stem  
40742  
Elbow, speed-fit, 1/4 x 1/4 stem  
40569  
40570  
40601  
40605  
Elbow, speed-fit, 3/8 x 1/4 mnpt, swivel  
Elbow, speed-fit, 1/4 x 1/4 mnpt, swivel  
Feed or Inlet Port, Quick-disconnect, (Female)  
Garden Hose Thread, female  
40604  
Garden Hose Thread, male  
42046  
40603  
42051  
Product & Waste Port, Quick-disconnect, (Female)  
Quick-disconnect (Male) for hoses  
Quick-disconnect (Male) for hoses, elbow  
CLEANERS  
1228653  
1228791  
Cleaner, Liqui-Clean403™,1gal., Low pH, liquid  
Cleaner, Liqui-Clean211™,1gal., High pH, liquid  
Membrane Cleaner, BIOSAN™ II, 1 gal., liquid  
14962  
TESTS  
50033  
Hardness Test Kit  
14604  
11656  
Silt Density Index (SDI) Test Kit  
Total Chlorine Test Kit  
70001  
Water Test Kit, AAMI Standards  
OPTIONAL  
14693e - 31 Mar 03  
7 - 4  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
12128  
40694  
40695  
40696  
40698  
14924  
Pre-filter Housing Assembly  
O-ring, Pre-filter (9 7/8”)  
Filter, 1 Micron, 9 7/8"  
Filter, 5 Micron, 9 7/8"  
Filter, Carbon, 9 7/8"  
Kit, Mil, Prefilter Install  
(Mounts prefilter to RO or to wall. Includes all necessary  
parts.)  
14693e - 31 Mar 03  
7 - 5  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Millenium™ RO System by GE Osmonics  
NOTES  
14693e - 31 Mar 03  
7 - 6  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
C a ll (253) 395-2200 for a dditiona l informa tion, (800) 633-3080 in the U.S., or visit www.osmonics.com  
Corporate Headquarters  
5951 C lea rwa ter Drive  
Minnetonka , MN  
55343-8995  
Kent Operations  
nd  
7848 South 202 ST  
Kent, WA  
98032-1345  
USA  
USA  
(952) 933 2277 Phone  
(952) 933-0141 Fa x  
(253) 395-2200 Phone  
(253) 395-2363 Fa x  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Dynex Video Gaming Accessories DX C114200 User Manual
Everpure Water System S 104 User Manual
Fisher Paykel Sleep Apnea Machine HC230 User Manual
Fortinet Network Router 100 User Manual
Fujitsu Laptop N6110 User Manual
Fujitsu Server TX150 S3 User Manual
Fusion Car Speaker PP CM525 User Manual
GE Portable Radio Sedona Series User Manual
GE Toaster 169211 User Manual
Global Machinery Company Drill 183T User Manual