Fluke Thermometer ESA612 User Manual

ESA612  
Electrical Safety Analyzer  
Users Manual  
March 2009, Rev. 1  
© 2009 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.  
All product names are trademarks of their respective companies.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents (continued)  
Notices  
All Rights Reserved  
© Copyright 2008, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or  
translated into any language without the written permission of Fluke Biomedical.  
Copyright Release  
Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service  
training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke  
Biomedical.  
Unpacking and Inspection  
Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking  
the instrument. Notify the carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking  
instructions, but be careful not to damage the instrument when unpacking it. Inspect the instrument for physical damage such as bent or  
broken parts, dents, or scratches.  
Technical Support  
For application support or answers to technical questions, either email [email protected] or call 1-800- 648-7952 or 1-425-  
446-6945.  
Claims  
Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing materials in  
their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not  
operate within specifications, or if there are any other problems not caused by shipping damage, please contact Fluke Biomedical or your local  
sales representative.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Standard Terms and Conditions  
Refunds and Credits  
Please note that only serialized products and their accessory items (i.e., products and items bearing a distinct serial number tag) are  
eligible for partial refund and/or credit. Nonserialized parts and accessory items (e.g., cables, carrying cases, auxiliary modules, etc.)  
are not eligible for return or refund. Only products returned within 90 days from the date of original purchase are eligible for refund/credit. In  
order to receive a partial refund/credit of a product purchase price on a serialized product, the product must not have been damaged by the  
customer or by the carrier chosen by the customer to return the goods, and the product must be returned complete (meaning with all manuals,  
cables, accessories, etc.) and in “as new” and resalable condition. Products not returned within 90 days of purchase, or products which are not  
in “as new” and resalable condition, are not eligible for credit return and will be returned to the customer. The Return Procedure (see below)  
must be followed to assure prompt refund/credit.  
Restocking Charges  
Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess  
of 30 days after purchase, but prior to 90 days, are subject to a minimum restocking fee of 20 %. Additional charges for damage and/or  
missing parts and accessories will be applied to all returns.  
Return Procedure  
All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory location. When you return an  
instrument to Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that  
you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are  
received in damaged condition due to improper packaging or handling.  
Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging:  
ƒ
ƒ
ƒ
Use a double–walled carton of sufficient strength for the weight being shipped.  
Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all projecting parts.  
Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents (continued)  
Returns for partial refund/credit:  
Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from our Order  
Entry Group at 1-800-648-7952 or 1-425-446-6945.  
Repair and calibration:  
To find the nearest service center, go to www.flukebiomedical.com/service or  
In the U.S.A.:  
Cleveland Calibration Lab  
Tel: 1-800-850-4606  
Everett Calibration Lab  
Tel: 1-888-99 FLUKE (1-888-993-5853)  
In Europe, Middle East, and Africa:  
Eindhoven Calibration Lab  
Tel: +31-402-675300  
In Asia:  
Everett Calibration Lab  
Tel: +425-446-6945  
Certification  
This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was  
shipped from the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for  
which there are no NIST calibration standards are measured against in-house performance standards using accepted test procedures.  
WARNING  
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper  
operation. Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment modifications.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Restrictions and Liabilities  
Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the  
information in this document will be incorporated in new editions of the publication. No responsibility is assumed by Fluke Biomedical  
for the use or reliability of software or equipment that is not supplied by Fluke Biomedical, or by its affiliated dealers.  
Manufacturing Location  
The ESA612 Electrical Safety Analyzer is manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Title  
Introduction .................................................................................................................... 1  
Safety Information .......................................................................................................... 2  
Intended Use.................................................................................................................. 3  
Unpacking the Analyzer ................................................................................................. 4  
Instrument Familiarization .............................................................................................. 4  
Connecting to Line Power .............................................................................................. 8  
Connecting a DUT to the Analyzer................................................................................. 8  
Turning the Analyzer On ................................................................................................ 8  
Accessing the Analyzer’s Functions............................................................................... 10  
Setting Up the Analyzer.................................................................................................. 11  
Setting Polarity Switching Delay................................................................................ 11  
Setting the Display Contrast...................................................................................... 11  
Setting up the Beeper................................................................................................ 12  
Viewing Instrument Information................................................................................. 12  
Viewing Memory........................................................................................................ 12  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
ESA612  
Users Manual  
Setting the GFCI Limit............................................................................................... 12  
Performing Electrical Safety Tests................................................................................. 12  
Setting the Test Standard ......................................................................................... 13  
Performing Mains Voltage Testing............................................................................ 13  
Performing a Ground Wire (Protective Earth) Resistance Test................................. 13  
Performing an Insulation Resistance Test ................................................................ 19  
Performing a Current Consumption Test .................................................................. 25  
Performing Leakage Current Tests........................................................................... 25  
Measuring Earth Leakage Current ....................................................................... 26  
Performing a Chassis (Enclosure) Leakage Test................................................. 29  
Performing a Lead-to-Ground (Patient) Leakage Test ......................................... 31  
Performing Lead-to-Lead (Patient Auxiliary) Leakage Tests................................ 33  
Performing a Lead Isolation (Mains on Applied Part) Leakage Test ......................... 35  
Performing an Alternative Equipment Leakage Test................................................. 37  
Performing an Alternative Applied Part Leakage Test .............................................. 37  
Performing a Direct Equipment Leakage Test .......................................................... 39  
Performing a Direct Applied Part Leakage Test........................................................ 41  
Performing a Differential Leakage Current Test........................................................ 44  
Using the 1210 Adapter................................................................................................. 46  
Making Point-To-Point Measurements........................................................................... 50  
Measuring Voltage.................................................................................................... 50  
Measuring Resistance .............................................................................................. 50  
Measuring Current .................................................................................................... 51  
Simulating ECG Waveforms.......................................................................................... 51  
Using Memory ............................................................................................................... 54  
Storing Data into Memory ......................................................................................... 54  
Viewing Memory Data............................................................................................... 55  
Deleting Data from Memory...................................................................................... 55  
Controlling the Analyzer Remotely................................................................................. 55  
ii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents (continued)  
Maintenance................................................................................................................... 56  
Testing and Replacing the Fuses................................................................................... 56  
Cleaning the Analyzer .................................................................................................... 57  
Replaceable Parts.......................................................................................................... 58  
Accessories.................................................................................................................... 60  
Specifications................................................................................................................. 61  
Detailed Specifications................................................................................................... 62  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ESA612  
Users Manual  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
List of Tables  
Table  
Title  
Page  
1.  
2.  
3.  
4.  
5.  
6.  
7.  
Symbols................................................................................................................................. 2  
Top-Panel Controls and Connections.................................................................................... 5  
Side and Top-Panel Connections.......................................................................................... 7  
Schematic Abbreviations....................................................................................................... 17  
Test Names Based on Selected Standard............................................................................. 25  
Replaceable Parts................................................................................................................. 58  
Accessories........................................................................................................................... 60  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
ESA612  
Users Manual  
vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure  
Title  
1.  
2.  
3.  
4.  
5.  
6.  
7.  
8.  
9.  
Front-Panel Controls and Connections ................................................................................. 4  
Side and Top-Panel Connections.......................................................................................... 6  
Analyzer Ready for Operation ............................................................................................... 8  
DUT Connections to the Analyzer ......................................................................................... 9  
Leakage Current Menu.......................................................................................................... 10  
Setup Menu........................................................................................................................... 11  
Mains Voltage Test Menu...................................................................................................... 13  
DUT Ground Resistance Measurement................................................................................. 14  
Ground Wire (Protective Earth) Resistance Measurement Connections............................... 16  
10. Ground Wire (Protective Earth) Resistance Measurement Schematic.................................. 18  
11. Insulation Resistance Measurement ..................................................................................... 19  
12. Mains to Protective-Earth Insulation Resistance Test Schematic.......................................... 20  
13. Applied Parts to Protective-Earth Insulation Test Schematic................................................. 21  
14. Mains to Applied Parts Insulation Test Schematic................................................................. 22  
15. Mains to Non-Earth Accessible Conductive Points Schematic.............................................. 23  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ESA612  
Users Manual  
16. Applied Parts to Non-Earth Conductive Points Schematic.................................................... 24  
17. Leakage Current Main Menu ................................................................................................ 26  
18. Earth Leakage Current Test Schematic................................................................................ 28  
19. Enclosure Leakage Current Test Schematic......................................................................... 30  
20. Lead-to-Ground (Patient) Leakage Current Test Schematic................................................. 32  
21. Applied Parts Connection Posts Display............................................................................... 33  
22. Lead-to-Lead (Patient Auxiliary) Leakage Current Test Schematic ...................................... 34  
23. Lead Isolation (Mains On Applied Parts) Leakage Test Schematic ...................................... 36  
24. Alternative Equipment Leakage Current Test Schematic...................................................... 38  
25. Alternative Applied Part Leakage Test Schematic................................................................ 40  
26. Direct Equipment Leakage Test Schematic.......................................................................... 42  
27. Direct Applied Parts Leakage Current Test Schematic......................................................... 43  
28. Differential Leakage Current Test Schematic ....................................................................... 45  
29. 1210 Adapter Connections ................................................................................................... 47  
30. EGC Lead Connection with 1210 Adapter ............................................................................ 49  
31. Point-To-Point Function Menu .............................................................................................. 50  
32. ECG Waveform Simulation Menu......................................................................................... 51  
33. ECG Monitor Connections .................................................................................................... 53  
34. Test Record ID Entry Screen................................................................................................ 54  
35. Fuse Access ......................................................................................................................... 57  
viii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Electrical Safety Analyzer  
Insulation resistance  
Introduction  
Ground (Earth) leakage  
Chassis (Enclosure) leakage  
The Fluke Biomedical ESA612 Electrical Safety Analyzer  
(hereafter the Analyzer) is a full-featured, compact,  
portable analyzer, designed to verify the electrical safety  
of medical devices. The Analyzer tests to domestic  
(ANSI/AAMI ES1, NFPA 99) and international (IEC62353,  
AN/NZS 3551, and parts of IEC 60601-1) electrical-safety  
standards. The integrated ANSI/AAMI ES1 and  
Lead to Ground (Patient) and Lead to Lead (Patient  
Auxiliary) leakage  
Lead isolation (Mains on applied parts leakage)  
Differential leakage  
Direct equipment leakage  
IEC60601-1 patient loads are easily selectable.  
Direct applied part leakage  
The Analyzer performs the following tests:  
Alternative equipment leakage  
Line (Mains) voltage  
Alternative applied part patient leakage  
Point to point leakage, voltage, and resistance  
ECG simulation and performance waveforms  
Ground Wire (or Protective Earth) Resistance  
Equipment current  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
Safety Information  
Table 1. Symbols  
Description  
In this manual, a Warning identifies hazardous conditions  
and actions that could cause bodily harm or death. A  
Caution identifies conditions and actions that could  
damage the Analyzer, the equipment under test, or cause  
permanent loss of data.  
Symbol  
W
Important information; refer to manual.  
Hazardous voltage  
X
XW Warning  
Conforms to relevant Canadian and US  
standards  
)
To avoid possible electrical shock or  
personal injury, follow these guidelines:  
Conforms to relevant Australian EMC  
requirements  
Use this Analyzer only in the manner  
specified by the manufacturer or the  
protection provided may be impaired.  
;
P
Conforms to European Union directives  
Read the Users Manual before operating  
Do not dispose of this product as  
unsorted municipal waste. Go to Fluke’s  
website for recycling information.  
the Analyzer.  
~
Do not connect the Analyzer to a patient or  
equipment connected to a patient. The  
Analyzer is intended for equipment  
evaluation only and should never be used  
in diagnostics, treatment or in any other  
capacity where the Analyzer would come  
in contact with a patient.  
IEC Measurement Category II – CAT II  
equipment designed to protect against  
transients from energy-consuming  
equipment supplied from fixed  
installations.  
CAT II  
Do not use the product in wet or damp  
Accessible Functional Earth Terminal  
˜
locations, around explosive gases or dust.  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Intended Use  
Inspect the Analyzer before using it. Do  
not use the Analyzer if abnormal  
Use extreme caution when working with  
voltages above 30 V.  
conditions of any sort are noted (such as a  
faulty display, broken case, etc.)  
Use the proper terminals, functions and  
ranges for the test being performed.  
Inspect the test leads for damaged  
insulation or exposed metal. Check test  
lead continuity. Replace damaged leads  
before using the Analyzer.  
Do not touch metal parts of the device  
under test (DUT) during analysis. The DUT  
should be considered an electrical shock  
hazard when connected to the Analyzer as  
some tests involve high voltages, high  
currents, and/or the removal of DUT earth  
bond.  
When testing, always be sure to keep your  
fingers behind the safety barriers on the  
test leads.  
Never open the Analyzer's case.  
Intended Use  
Dangerous voltages are present. There are  
no user replaceable parts in the Analyzer.  
The Analyzer is intended for use by trained service  
technicians to perform periodic inspections on a wide  
range of medical equipment. The testing procedures are  
menu-driven, and simple to operate.  
Have the Analyzer serviced only by  
qualified personnel.  
The Analyzer must be properly earthed.  
Only use a supply socket that has a  
protective earth contact. If there is any  
doubt as to the effectiveness of the supply  
socket earth, do not connect the Analyzer.  
Do not use a two-conductor adapter or  
extension cord; this will break the  
protective ground connection.  
Do not use the 15-20 A adapter to power  
devices rated in excess of 15 A. Doing so  
may overload the installation.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
Unpacking the Analyzer  
Instrument Familiarization  
Carefully unpack all items from the box and check that  
you have the following items:  
Figure 1 and Table 2 describes the front-panel controls  
and connections of the Analyzer.  
ESA612  
Getting Started Manual  
Users Manual CD  
Carrying case  
1
9
Power cord  
15 – 20 A Adapter (USA only)  
2
ESA USA Accessory Kit (USA, Australia, and Israel  
only)  
ESA EUR Accessory Kit  
Ansur demo CD  
8
3
Null Post Adapter  
5-to-5 Banana to ECG Adapter (BJ2ECG)  
Transfer cable  
4
TEST  
7
5
6
fis116.eps  
Figure 1. Front-Panel Controls and Connections  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Instrument Familiarization  
Table 2. Top-Panel Controls and Connections  
Item  
Name  
Description  
Item  
Name  
Description  
Connection for zeroing test  
lead resistance.  
Nulling Jack  
8
Controls the configuration of  
the equipment outlet. Opens  
and closes the neutral and  
ground connection and  
Equipment Outlet  
Configuration  
Buttons  
Keys F1 through F4 are used  
to select from a number of  
1
reverses the polarity of the  
neutral and hot connection.  
Function Softkeys selections that appear in the  
LCD display above each  
9
function softkey.  
Indicates when high voltage  
is applied to the  
High Voltage  
Indicator  
ECG/Applied Parts posts or  
L1 and L2 of the Test  
Receptacle.  
2
3
Test Function  
Buttons  
Selects the various Analyzer  
test functions.  
Cursor control buttons for  
navigating menus and lists.  
Navigation Buttons  
Test Button  
4
5
6
7
Initiates selected tests.  
Saves the measurement or  
ECG waveform to memory.  
Save Button  
Test lead connectors.  
Input Jacks  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
Figure 2 and Table 3 describe the side and top-panel connections of the Analyzer.  
RA  
LL  
F
L
N
LA  
RL  
1  
C1  
TEST  
7
2
4
1
3
8
5
6
fis110.eps  
Figure 2. Side and Top-Panel Connections  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Instrument Familiarization  
Table 3. Side and Top-Panel Connections  
Description  
Item  
Name  
Equipment Outlet  
Equipment outlet, specific to the version of the Analyzer, which provides a DUT  
connection.  
1
Digital connection for controlling the Analyzer from a PC or instrument  
controller.  
USB Device Port  
2
(B-style connector)  
3
4
5
Fuse Access Door  
Tilt Stand  
Covers the equipment outlet fuses.  
Support for holding the Analyzer in a tilted position.  
Turns ac power on and off.  
AC Power Switch  
A grounded male three-prong (IEC 60320 C19) connector that accepts the  
line-power cord.  
6
Power Input Connector  
Connection posts for Device Under Test (DUT) applied parts, such as ECG  
leads. Used to test for leakage current through leads and to supply ECG  
signals and performance waveforms to a DUT.  
7
8
ECG/Applied Parts Jacks  
Banana Jack to ECG Adapter  
Adapter for connecting ECG snap leads to the Analyzer.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
Turning the Analyzer On  
Connecting to Line Power  
Note  
XWWarning  
To ensure the high voltage indicator is working,  
look for it to illuminate during the power-up self  
test.  
To avoid shock hazard and for proper  
Analyzer operation, connect the factory  
supplied three-conductor line power cord to  
a properly grounded power outlet. Do not  
use a two-conductor adapter or extension  
cord; this will break the protective ground  
connection.  
Press the power switch on the left-side panel so the “I”  
side of the ac power switch is depressed. The Analyzer  
will perform a series of self tests and then display the  
message shown in Figure 3 when the self test has  
completed successfully.  
Connect the Analyzer to a properly grounded three-prong  
outlet. The Analyzer will not properly test a DUT when the  
ground lead is open.  
The Analyzer is intended for use with single-phase,  
grounded power. It is not intended for dual, split-phase or  
three-phase power configurations. But it can be used with  
any power system that supplies the correct voltages for  
single-phase and is grounded, or is an isolated power  
system.  
Connecting a DUT to the Analyzer  
A Device Under Test (DUT) can be connected in a  
number of different ways depending on the device and  
the number of connections needed for a full electrical  
safety test. Figure 4 shows a DUT connected to the test  
receptacle, applied parts posts, and a separate  
connection to the DUT’s enclosure or protective earth  
ground.  
fis101.jpg  
Figure 3. Analyzer Ready for Operation  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Turning the Analyzer On  
To protective earth or any exposed  
conductive surface on the enclosure  
A  
R
L  
LA  
L
L  
C
1  
1
TEST  
fis113.eps  
Figure 4. DUT Connections to the Analyzer  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
These icons indicate the use of E and F to select an  
applied part.  
During the self-test, the Analyzer checks its ac mains  
input for proper polarity, ground integrity and voltage  
level. The high voltage indicator illuminates briefly during  
the self test. If the polarity is reversed, the Analyzer  
indicates this condition and allows the polarity to be  
reversed internally. If the ground is open, the Analyzer  
displays this fault. If the mains voltage is too high or too  
low, the Analyzer displays this fault and does not continue  
until the supply voltage is corrected and the Analyzer  
power cycled off and then on again.  
Accessing the Analyzer’s Functions  
For each test and setup function, the Analyzer uses a  
series of menus to access various Analyzer test and  
setup variables. As shown in Figure 5, the Analyzer  
indicates various leakage current tests along the bottom  
of the display. An Exit selection is also indicated as a way  
of backing out of the leakage current tests. Pressing a  
softkey (F1 through F4) under a specific test will cause  
the analyzer to setup for or perform the selected test.  
fis102.jpg  
Figure 5. Leakage Current Menu  
The three buttons along the right side of the display  
(PND) control the wiring of the Analyzer’s  
test receptacle for some electrical tests. The present state  
of these three buttons is displayed along the right edge of  
the display whenever these controls are active.  
In addition to the function softkeys, the Analyzer test  
functions may require using the navigation buttons to  
select parameters as well. In the example above, the  
leakage selection has K next to it. This icon indicates the  
selection is controlled by pressing G or H. In this  
example, the leakage current measurement is switched  
between AC+DC, AC only, or DC only. The applied parts  
indicator has W on the left end and X on the right end.  
Figure 5 shows polarity is settable between normal,  
reversed, and off. Neutral is also settable to closed or  
open. Earth condition is not displayed, which indicates it  
can not be changed. However, earth is internally opened  
during this test.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Setting Up the Analyzer  
Setting Polarity Switching Delay  
Setting Up the Analyzer  
When switching the polarity of the Analyzer’s test  
receptacle, a delay can be set to control the actual switch  
time. To set the polarity delay:  
There are a number of Analyzer parameters that are  
adjusted through a setup function as well as the ability to  
save a record by ID and date. To access the first Setup  
menu shown in Figure 6, press Q.  
1. From the setup menu, press the softkey labeled  
More to reveal additional menu selections.  
2. Press the softkey labeled Instrument to reveal the  
instrument setup selections.  
3. Press the softkey labeled Polarity Delay to open the  
scroll box above the softkey label.  
4. Press G or H to adjust the delay from 0 to 5  
seconds in 1 second steps.  
5. Press the softkey labeled Done to exit the switching  
polarity delay setup function.  
Setting the Display Contrast  
fis114.bmp  
There are two methods for setting the display contrast.  
From the Select a Test…. menu or through the setup  
menu.  
Figure 6. Setup Menu  
Note  
See the Using Memory section later in this  
manual for a description of how to enter a test  
record ID.  
Whenever the Analyzer displays its start-up menu (Select  
a test…), pressing G or H will increase or decrease the  
display’s contrast respectively. Press the softkey labeled  
Done to exit contrast setup.  
The setup parameters have been grouped into six  
categories: Instrument, Display, Sound, Instrument Info,  
Calibration, and Diagnostics.  
Another way to adjust the contrast is through the  
Analyzer’s setup menu.  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
1. From the setup menu, press the softkey labeled  
Viewing Memory  
More twice to set F1 to the display contrast function.  
Refer to the Using Memory section later in the manual to  
learn about viewing memory and how to store data in the  
Analyzer.  
2. Press the softkey labeled Display Contrast.  
3. Press G or H to increase or decrease the display’s  
contrast respectively.  
Setting the GFCI Limit  
4. Press the softkey labeled Done to exit contrast  
To set the GFCI current limit:  
setup.  
1. From the setup menu, press the softkey labeled  
Setting up the Beeper  
More to reveal additional menu selections.  
To enable or disable the beeper:  
2. Press the softkey labeled Instrument to reveal the  
instrument setup selections.  
1. From the setup menu, press the softkey labeled  
More twice to set F2 to the beeper on/off function.  
3. Press the softkey labeled GFCI Limit to open the  
scroll box above the softkey label.  
2. Press the softkey labeled Beeper to switch the  
beeper on and off.  
4. Press G or H to adjust the current limit from 5 to  
25 mA.  
3. Press the softkey labeled Done to go back to the  
setup menu.  
5. Press the softkey labeled Done to exit the GFCI Limit  
setup function.  
Viewing Instrument Information  
Performing Electrical Safety Tests  
To view information about the Analyzer:  
The Analyzer is designed to perform a number of different  
electrical and performance tests on biomedical  
equipment. The following sections describe the various  
tests and how to perform them using the Analyzer.  
1. From the setup menu, press the softkey labeled  
More twice to set F3 to the instrument information  
function.  
2. Press the softkey labeled Instrument Information.  
3. After viewing the displayed information, press the  
softkey labeled Done to exit the information screen.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
Setting the Test Standard  
The Analyzer is designed to perform electrical safety  
testing based on a number of different safety standards:  
AAMI ES1/NFPA99, IEC62353, IEC60601-1, and  
AN/NZS 3551. AAMI is the Analyzer’s default standard.  
To select another standard:  
1. Press Q.  
2. From the setup menu, press the softkey labeled  
More to reveal additional menu selections.  
3. Press the softkey labeled Instrument to reveal the  
fis104.jpg  
instrument setup selections.  
Figure 7. Mains Voltage Test Menu  
4. Press the softkey labeled Standard to open the  
Press each function softkey to perform each of the three  
measurements: Live to neutral, neutral to earth, and live  
to earth.  
scroll box above the softkey label.  
5. Press G or H to scroll through the standard  
selections.  
Note  
6. When the desired standard is displayed, press the  
Power to the test receptacle is off during the  
Mains Voltage test.  
softkey labeled Done.  
Some electrical tests may not be applicable for a specific  
standard. In these cases, the Analyzer’s menu will not  
display the excluded test as a selection.  
Performing a Ground Wire (Protective Earth)  
Resistance Test  
The Ground Wire (Protective-Earth) Resistance test  
measures the impedance between the Analyzer’s test  
receptacle’s PE terminal and the exposed conductive  
parts of the DUT that are connected to the DUT’s  
Protective Earth.  
Performing Mains Voltage Testing  
The Mains Voltage test measures the voltage on the  
mains input through three separate measurements. To  
access the Mains Voltage test, press V. The Mains  
Voltage test menu is shown in Figure 7.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
Prior to conducting any leakage tests with the Analyzer, it  
is best to test the integrity of the ground connection  
between the Analyzer’s test receptacle ground and the  
DUT’s Protective earth ground or enclosure with this test.  
the DUT enclosure or protective earth connection.  
8. The measured resistance is displayed as shown in  
Figure 8 after the DUT connection(s) is/are made.  
To access the Ground Wire (Protective Earth) /Null  
Resistance Test menu press O.  
Note  
The DUT is powered off for this test.  
To perform a ground wire resistance test:  
1. Ensure the power cord from the DUT is plugged into  
the Analyzer’s test receptacle.  
2. Press O to reveal the resistance function menu.  
3. Connect one end of a test lead to the V/Ω/A jack as  
fis105.jpg  
shown in Figure 9.  
Figure 8. DUT Ground Resistance Measurement  
4. If using an accessories probe, connect it to the other  
end of the test lead and place the probe tip into the  
/Null jack. If using an alligator clip accessory,  
connect it to the other end of the test lead, place the  
null post adapter in the /Null jack, and clamp the  
alligator clip to the null post adapter.  
WX Warning  
To avoid electric shock, remove the null post  
adapter from the /Null jack after a test lead  
zero is performed. The /Null jack becomes  
potentially hazardous during some of the test  
conditions.  
5. Connect the other end of the test lead to /Null jack.  
A low resistance reading is required to confirm a good  
ground connection through the power cord. Refer to the  
appropriate electrical safety standard for the specific limit  
value to be followed.  
6. Press the softkey labeled Zero Leads. The Analyzer  
zeroes out the measurement to cancel the test lead  
resistance.  
7. Connect the test lead coming from the /Null jack to  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
Figure 9 shows the electrical connections between the  
Analyzer and the DUT. Table 4 lists the abbreviations  
used in the schematics and their descriptions.  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ESA612  
Users Manual  
To protective earth or any exposed  
conductive surface on the enclosure  
RA  
F
LL  
RL  
N
C
1
TEST  
fis112.eps  
Figure 9. Ground Wire (Protective Earth) Resistance Measurement Connections  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
Table 4. Schematic Abbreviations  
Abbreviation  
Meaning  
MD  
Measuring Device (ESA612 Analyzer)  
FE  
Functional Earth  
Protective Earth  
PE  
Mains  
L1  
Mains Voltage Supply  
Live Conductor  
L2  
Neutral Conductor  
Device Under Test  
DUT  
DUT_L1  
DUT_L2  
DUT_PE  
REV POL  
LEAD GND  
MAP  
Device Under Test Live conductor  
Device Under Test neutral conductor  
Device Under Test protective earth  
Reversed mains supply polarity  
Lead to ground, used in Patient leakage test  
Mains on Applied Part  
MAP REV  
PE Open  
Reverse mains on applied part source voltage  
Open protective earth  
Test Voltage  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
DEVICE UNDER TEST  
DUT_L1  
DUT_L2  
APPLIED  
PART  
EARTH  
OPEN  
FE  
DUT_PE  
PE  
CONDUCTIVE PART  
MD  
TEST LEAD  
faw26.eps  
Figure 10. Ground Wire (Protective Earth) Resistance Measurement Schematic  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
Performing an Insulation Resistance Test  
The five insulation resistance tests take measurements on  
mains (L1 & L2) to Protective earth, applied parts to  
Protective earth, mains to applied parts, mains to non-  
earthed accessible conductive points, and applied parts to  
non-earthed accessible conductive points.  
To access the Insulation Resistance Test menu,  
press M.  
All Insulation Resistance Tests can be performed using  
500 or 250 volts dc. To change the test voltage from the  
Insulation Resistance Test menu, press the softkey  
labeled More. Pressing the softkey labeled Change  
Voltage will cause the test voltage to toggle between 250  
and 500 volts dc.  
fis106.jpg  
Figure 11. Insulation Resistance Measurement  
After selecting one of the tests by pressing the appropriate  
softkey, press T to apply the selected voltage to the  
DUT and take the resistance measurement.  
Note  
Exiting and re-entering the Insulation Resistance  
Test menu causes the test voltage to return to its  
default value of 500 volts dc.  
Figures 12 through 16 shows the electrical connections  
between the Analyzer and DUT for the five insulation  
resistance tests.  
As shown in Figure 11, three of the five tests are shown  
over function soft keys F1 through F3. To access the other  
two tests or test voltage selection, press the softkey  
labeled More. The softkey labeled Back will move the  
menu back up to the top-level insulation resistance test  
menu.  
Note  
The DUT is powered off for this test.  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
DEVICE UNDER TEST  
DUT_L1  
DUT_L2  
APPLIED  
PART  
MD  
DUT_PE  
FE  
CONDUCTIVE PART  
faw17.eps  
Figure 12. Mains to Protective-Earth Insulation Resistance Test Schematic  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
DEVICE UNDER TEST  
DUT_L1  
DUT_L2  
APPLIED  
PART  
DUT_PE  
FE  
CONDUCTIVE PART  
MD  
faw18.eps  
Figure 13. Applied Parts to Protective-Earth Insulation Test Schematic  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
DEVICE UNDER TEST  
DUT_L1  
DUT_L2  
APPLIED  
PART  
MD  
DUT_PE  
FE  
CONDUCTIVE PART  
faw19.eps  
Figure 14. Mains to Applied Parts Insulation Test Schematic  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
DEVICE UNDER TEST  
DUT_L1  
DUT_L2  
APPLIED  
PART  
MD  
DUT_PE  
FE  
CONDUCTIVE PART  
TEST  
LEAD  
faw20.eps  
Figure 15. Mains to Non-Earth Accessible Conductive Points Schematic  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
DEVICE UNDER TEST  
DUT_L1  
DUT_L2  
APPLIED  
PART  
DUT_PE  
FE  
CONDUCTIVE PART  
TEST  
LEAD  
MD  
faw21.eps  
Figure 16. Applied Parts to Non-Earth Conductive Points Schematic  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
Which leakage tests are available depends on which  
standard is selected. See the Selecting the Test Standard  
section earlier in this manual to change the standard the  
Analyzer is using.  
Performing a Current Consumption Test  
To measure the current consumed by the DUT, press  
A. The Analyzer displays the current flowing through  
the mains connections of the test receptacle.  
Table 5 lists six leakage current tests that have different  
names based on which standard is selected.  
Performing Leakage Current Tests  
The Analyzer measures leakage current for a number of  
different DUT configurations. In addition to the leakage  
found on the enclosure and the earth connection, the  
Analyzer can measure leakage on each connected applied  
part and combinations of connected applied parts.  
Press B to access the leakage current main menu  
shown in Figure 17.  
Table 5. Test Names Based on Selected Standard  
AAMI/NFPA 99  
IEC60601  
Protective Earth Resistance  
Ground Wire Resistance  
Earth Leakage Current  
Ground Wire Leakage Current  
Chassis Leakage Current  
Touch or Enclosure Leakage Current  
Patient Leakage Current  
Lead to Ground Leakage Current  
Lead to Lead Leakage Current  
Isolation Leakage Current  
Patient Auxiliary Leakage Current  
Mains on Applied Part (MAP) Leakage Current  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
Measuring Earth Leakage Current  
Note  
The Ground Wire (Earth) Leakage test is  
available for AAMI, 60601, and not IEC 62353.  
To measure the current flowing in the DUT’s protective  
earth circuit, press the softkey labeled Ground Wire  
(pending the standard) from the leakage current main  
menu. Figure 18 shows the electrical connections between  
the Analyzer and the DUT during a Ground Wire Leakage  
Current Test.  
Within the Ground Wire Leakage Current test there are a  
few combination measurements that can be performed.  
Pressing P switches the polarity of the mains voltage  
applied to the Analyzer’s test receptacle between Normal,  
Off, Reverse, and Off. Pressing N opens and closes  
the neutral connection to the Analyzer’s test receptacle.  
There is no need to open up the test receptacle earth  
(ground), since this is done internally during the  
measurement.  
fis117.jpg  
Figure 17. Leakage Current Main Menu  
Note  
The display shown in Figure 17 is the main  
leakage current menu when AAMI is the selected  
standard.  
All leakage currents, with the exception of Lead Isolation  
(Mains on Applied parts), are displayed in one of three  
ways: AC+DC, AC Only, or DC only. The initial result is  
displayed in the appropriate parameter based on the  
standard selected. To change the displayed parameter,  
press G or H. The present measurement method is  
displayed to the right of the current measurement while  
leakage current tests are conducted.  
The following outlet conditions apply when performing this  
test:  
Normal Polarity  
Normal Polarity, Open Neutral  
Reversed Polarity  
Reversed Polarity, Open Neutral  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
IEC60601-1 specifies that the applied parts should be  
connected for this measurement. Enable this  
measurement by pressing E or F which grounds and  
ungrounds all applied parts connection posts.  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ESA612  
Users Manual  
DEVICE UNDER TEST  
MAINS  
ON  
DUT_L1  
DUT_L2  
L1  
MAINS  
L2  
APPLIED  
PART  
L2  
OPEN  
REV  
POL  
EARTH  
OPEN  
DUT_PE  
FE  
CONDUCTIVE PART  
MD  
PE  
APPLIED PART  
OPEN  
faw27.eps  
Figure 18. Earth Leakage Current Test Schematic  
Note  
Ground wire leakage is the same schematic without the Applied Parts switch.  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
Normal Polarity, Open Neutral  
Performing a Chassis (Enclosure) Leakage Test  
Note  
Reversed Polarity  
Reversed Polarity, Open Earth  
Reversed Polarity, Open Neutral  
The Chassis (Enclosure) Leakage test is only  
available for the IEC60601 or ANSI/AAMI ES1  
1993 standard selections.  
IEC60601-1 specifies that the applied parts should be  
connected for this measurement. Enable this  
measurement by pressing E or F which grounds and  
ungrounds all applied parts connection posts.  
The Chassis (Enclosure) Leakage Test measures the  
current flowing between the DUT’s enclosure and  
protective earth. Figure 19 shows the electrical  
connections between the Analyzer and the DUT.  
To perform a Chassis (Enclosure) Leakage Test:  
1. Connect a lead between the Analyzer’s V/Ω/A jack  
and the DUT’s enclosure.  
2. Press the softkey labeled Chassis from the Leakage  
Current Test menu.  
3. The Analyzer displays the measured current.  
The Chassis Leakage test can be performed with a  
number of fault conditions on the test receptacle. Press  
P to switch the test receptacle between Normal, Off,  
Reverse, and Off. Press N to open and close the  
neutral connection to the receptacle. Press D to open  
and close the receptacle’s earth connection.  
The following outlet conditions apply when performing this  
test:  
Normal Polarity  
Normal Polarity, Open Earth  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
DEVICE UNDER TEST  
MAINS  
ON  
DUT_L1  
DUT_L2  
L1  
MAINS  
L2  
APPLIED  
PART  
L2  
OPEN  
EARTH  
OPEN  
DUT_PE  
FE  
CONDUCTIVE PART  
TEST LEAD  
MD  
PE  
APPLIED PART  
OPEN  
faw28.eps  
Figure 19. Enclosure Leakage Current Test Schematic  
Note  
Chassis leakage is the same schematic without the Applied Parts switch.  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
The Lead-to-Ground Leakage test can be performed with  
a number of fault conditions on the test receptacle. Press  
P to switch the test receptacle between Normal, Off,  
Reverse, and Off. Press N to open and close the  
neutral connection to the receptacle. Press D to open  
and close the receptacle’s earth connection.  
Performing a Lead-to-Ground (Patient) Leakage  
Test  
Note  
The Lead-to-Ground (Patient) Leakage Current  
Test is not available for IEC 62353 standard  
selections.  
The following outlet conditions apply when performing this  
test:  
The Lead-to-Ground (Patient) Leakage Current test  
measures the current flowing between a selected applied  
part, selected group of applied parts, or ALL applied parts,  
and the Mains PE. Figure 20 shows the electrical  
connections between the Analyzer and the DUT.  
Normal Polarity  
Normal Polarity, Open Neutral  
Normal Polarity, Open Earth  
Reversed Polarity  
To perform a Lead-to Ground (Patient) leakage test:  
1. Press B.  
Reversed Polarity, Open Neutral  
Reversed Polarity, Open Earth  
2. Press the soft key labeled More.  
Note  
3. Select one of the applied part groupings by pressing  
If there are more than five applied parts to  
connect to the Analyzer, see Using the 1210  
Adapter later in this manual.  
G or H.  
Note  
Refer to the testing standard when deciding the  
type of the applied parts and how they should be  
grouped for testing.  
4. Press the soft key labeled Select.  
5. Press F or E to advance through each applied part  
grouping, or the individual applied parts, to ground.  
These are selected and measured.  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
DEVICE UNDER TEST  
MAINS  
ON  
DUT_L1  
DUT_L2  
L1  
APPLIED  
PART  
MAINS  
L2  
L2  
OPEN  
REV  
POL  
EARTH  
OPEN  
DUT_PE  
FE  
CONDUCTIVE PART  
LEAD SELECT  
MD  
PE  
LEAD GND  
faw29.eps  
Figure 20. Lead-to-Ground (Patient) Leakage Current Test Schematic  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
different combinations of posts tied together: RA/R and  
LL/F, RA/R and LA/L, or LL/F and LA/L.  
Performing Lead-to-Lead (Patient Auxiliary) Leakage  
Tests  
Note  
The Lead-to-Lead (Patient Auxiliary) leakage test  
is available when the IEC60601 or ANSI/AAMI  
ES1-1993 standard is selected.  
To measure the leakage current through each applied part  
or lead and selected combination of lead connections (all  
other or between two), press the softkey labeled Lead to  
Lead from the Leakage Test main menu shown in  
Figure 17. Figure 22 shows the electrical connections  
between the Analyzer and the DUT during a Lead-to-Lead  
(Patient Auxiliary) Leakage Current Test.  
fis107.eps  
Figure 21. Applied Parts Connection Posts Display  
Within the Lead-to-Lead (Patient Auxiliary) Leakage test, a  
number of fault measurements can be made. Pressing  
P switches the polarity of the mains voltage applied  
to the Analyzer’s test receptacle between Normal, Off,  
Reverse, and Off. Pressing N opens and closes the  
neutral connection to the Analyzer’s test receptacle.  
Pressing Dopens and closes the earth or ground  
connection to the Analyzer’s test receptacle.  
The Lead-to-Lead (Patient Auxiliary) Leakage test adds a  
diagram of the applied parts connection posts to the  
display, as shown in Figure 21. In the figure, the applied  
parts post RA/R is shown above the other posts. This  
indicates that the leakage measurement is being made  
from RA/R to all others. To move to the next applied part  
post, press F. The first post will appear inline with the  
other posts while the LL/F post appears above all others.  
This indicates the second leakage measurement is being  
made from LL/F to all others. Continue pressing F or E to  
move from one connection post to another and noting the  
measured current in the display.  
Note  
If there are more than five applied parts to  
connect to the Analyzer, see Using the 1210  
Adapter later in this manual.  
After each post is isolated individually, the Lead-to-Lead  
(Patient Auxiliary) Leakage test measures current of three  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
MAINS  
ON  
DEVICE UNDER TEST  
DUT_L1  
DUT_L2  
L1  
APPLIED  
PART  
MAINS  
L2  
L2  
OPEN  
REV  
POL  
EARTH  
OPEN  
DUT_PE  
FE  
CONDUCTIVE PART  
PE  
+ LEAD  
MD  
- LEAD  
GROUND LEAD  
faw30.eps  
Figure 22. Lead-to-Lead (Patient Auxiliary) Leakage Current Test Schematic  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
The following outlet conditions apply when performing this  
test:  
1. Press B.  
2. Press the soft key labeled More.  
Normal Polarity  
3. Select the desired applied part groupings using  
Normal Polarity, Open Neutral  
Normal Polarity, Open Earth  
Reversed Polarity, Open Neutral  
Reversed Polarity, Open Earth  
G and H.  
Note  
Refer to the testing standard when deciding the  
type of the applied parts and how they should be  
grouped for testing.  
Performing a Lead Isolation (Mains on Applied  
Part) Leakage Test  
4. Press the soft key labeled Select.  
Note  
5. Press the soft key labeled Lead Isolation.  
The Lead Isolation (Mains on Applied Part)  
leakage test is available when the IEC60601 &  
ANSI/AAMI standard is selected.  
6. Press F or E to select the desired applied part  
connection.  
7. Press T to apply the voltage and read the leakage  
The Lead Isolation (Mains On Applied Parts) Leakage  
Current test measures the current that flows in response  
to an isolated AC voltage applied between a selected  
applied part, group of applied parts, or ALL applied parts,  
and Earth (and any conductive part connected to the RED  
terminal). Figure 23 shows the electrical connections  
between the Analyzer and the DUT during a Mains on  
Applied Part Leakage Current Test.  
current in the display.  
Pressing E and F scrolls through the applied part  
connections or groupings. Press T for each  
connection configuration to thoroughly test the DUT.  
The following outlet conditions apply when performing this  
test:  
Normal Polarity  
Reverse Polarity  
Note  
Note  
With 60601 standard selected, the MAP test  
voltage is available in both Normal and Reverse  
(180 degrees out of phase with mains).  
If there are more than five applied parts to  
connect to the Analyzer, see Using the 1210  
Adapter later in this manual.  
To perform a Lead Isolation (Mains on Applied Part) test:  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
DEVICE UNDER TEST  
MAINS  
ON  
DUT_L1  
L1  
MAINS  
L2  
APPLIED  
PART  
L2  
OPEN  
DUT_L2  
DUT_PE  
REV  
POL  
EARTH OPEN  
FE  
CONDUCTIVE PART  
TEST LEAD  
PE  
MD  
LEAD SELECT  
MAP (ISOLATION)  
TRANSFORMER  
MAP  
REV  
faw31.eps  
Figure 23. Lead Isolation (Mains On Applied Parts) Leakage Test Schematic  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
The following outlet conditions apply when performing this  
test:  
Performing an Alternative Equipment Leakage  
Test  
Closed Earth  
Open Earth  
Note  
The alternative equipment leakage test is  
available when the EN62353 standard is  
selected.  
Note  
If there are more than five applied parts to  
connect to the Analyzer, see Using the 1210  
Adapter later in this manual.  
During the Alternative Equipment Leakage test, the  
voltage source is applied between short-circuited  
equipment outlet mains live, neutral, and equipment outlet  
earth, the exposed conductive surface on the housing, and  
all applied parts short-circuited together. Equipment is  
separated from mains during the test. The current which  
flows over the insulation of the DUT is measured.  
Performing an Alternative Applied Part Leakage  
Test  
Note  
The Alternative applied part leakage test is  
available when the EN62353 standard is  
selected.  
This test is not applicable for equipment with internal  
electrical power source. The switches in mains part shall  
be closed during measurement.  
During the Alternative Applied Part Leakage test, the test  
voltage is applied between short-circuited applied parts of  
a single function and the short-circuited equipment outlet  
mains live, neutral, equipment outlet earth, and exposed  
conductive surface on the housing. This test should only  
be done for equipment with F-Type applied parts. For  
equipment with multiple applied parts, test each group of  
applied parts of a single function in turn with all others  
floating during the test. All applied parts can be connected  
to the Analyzer’s applied parts jacks and the lead selection  
will float those not selected.  
To perform an alternative equipment leakage test:  
1. Press B.  
2. Press the softkey labeled Alternative Equipment.  
3. Press T to apply the voltage and read the current  
in the display.  
Figure 24 shows the electrical connections between the  
Analyzer and the DUT during an Alternative Equipment  
Leakage Test.  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
DEVICE UNDER TEST  
DUT_L1  
DUT_L2  
APPLIED  
PART  
DUT_PE  
FE  
CONDUCTIVE PART  
PE  
OPEN  
TEST  
LEAD  
MD  
faw22.eps  
Figure 24. Alternative Equipment Leakage Current Test Schematic  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
To perform an alternative applied part leakage test:  
1. Press B.  
The Direct Equipment Leakage Current test measures the  
leakage current between all applied parts and the exposed  
conductive surface on the housing, to mains earth.  
2. Press the soft key labeled More.  
To perform a direct equipment test:  
3. Select the desired applied part groupings using  
1. Press B.  
G and H.  
The direct equipment test is the default test and should  
already be selected.  
4. Press the soft key labeled Select.  
5. Press the soft key labeled Alternative A.P..  
2. Press T to apply the voltage and read the leakage  
6. Press T to apply the test voltage and read the  
current in the display  
current in the display.  
Figure 26 shows the electrical connections between the  
Analyzer and the DUT during a Direct Equipment Leakage  
Current Test.  
7. Press F or E to advance to the next applied part  
group(s) of a single function if applicable. Pressing  
T to read leakage current for each group.  
The following outlet conditions apply when performing this  
test:  
Figure 25 shows the electrical connections between the  
Analyzer and the DUT during an Alternative Applied Part  
Leakage current test.  
Normal Polarity, Closed Earth  
Normal Polarity, Open Earth  
Reversed Polarity, Closed Earth  
Reversed Polarity, Open Earth  
Note  
If there are more than five applied parts to  
connect to the Analyzer, see Using the 1210  
Adapter later in this manual.  
Note  
If there are more than five applied parts to  
connect to the Analyzer, see Using the 1210  
Adapter later in this manual.  
Performing a Direct Equipment Leakage Test  
Note  
The Direct Equipment Leakage test is available  
when the EN62353 standard is selected.  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
DEVICE UNDER TEST  
DUT_L1  
DUT_L2  
AP1  
AP2  
APPLIED  
PART  
DUT_PE  
FE  
CONDUCTIVE PART  
TEST  
LEAD  
MD  
faw23.eps  
Figure 25. Alternative Applied Part Leakage Test Schematic  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
7. Press F or E to advance to the next group of applied  
Performing a Direct Applied Part Leakage Test  
parts, if applicable.  
Note  
Figure 27 shows the electrical connections between the  
Analyzer and the DUT during a Direct Applied Part  
Leakage Current Test.  
The Direct Applied Part Leakage test is available  
when the EN62353 standard is selected.  
The Direct Applied Part Leakage Current test measures  
the leakage current between all applied parts of a single  
function and the exposed conductive surface on the  
housing, to mains earth. For equipment with multiple  
applied parts, each group of a single function should be  
tested each in turn with all other floating during the test.  
This test should only be done for equipment with F-Type  
applied parts.  
The following outlet conditions apply when performing this  
test:  
Normal Polarity  
Reversed Polarity  
Note  
For Type B applied part, see direct equipment leakage  
schematic in Figure 26.  
If there are more than five applied parts to  
connect to the Analyzer, see Using the 1210  
Adapter later in this manual.  
To perform a direct applied part leakage test:  
1. Press B.  
2. Press the soft key labeled More.  
3. Select the desired applied part groupings using  
G and H.  
4. Press the soft key labeled Select. The Direct A.P. test  
should already be selected.  
5. Press F or E to select the applied part test  
configuration.  
6. Press T to apply the test voltage and read the  
current in the display.  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
DEVICE UNDER TEST  
MAINS  
ON  
DUT_L1  
DUT_L2  
L1  
MAINS  
L2  
APPLIED  
PART  
L2  
OPEN  
REV  
POL  
PE  
OPEN  
DUT_PE  
FE  
CONDUCTIVE PART  
MD  
TEST  
LEAD  
faw24.eps  
Figure 26. Direct Equipment Leakage Test Schematic  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
DEVICE UNDER TEST  
MAINS  
ON  
DUT_L1  
DUT_L2  
AP1  
L1  
MAINS  
L2  
APPLIED  
PART  
AP2  
L2  
OPEN  
REV  
POL  
DUT_PE  
FE  
CONDUCTIVE PART  
MD  
faw25.eps  
Figure 27. Direct Applied Parts Leakage Current Test Schematic  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
Note  
Performing a Differential Leakage Current Test  
If there are more than five applied parts to  
connect to the Analyzer, see Using the 1210  
Adapter later in this manual.  
Note  
The Differential Leakage Current test is available  
when the EN62353 standard is selected.  
The differential leakage current test measures the  
magnitudes of the differential current flowing in the  
Equipment Outlet live and neutral, with power applied to  
the equipment outlet. All applied parts should be  
connected during this test, if equipment has applicable  
applied parts.  
To perform a differential leakage current test:  
1. Press B.  
2. Press the soft key labeled Differential.  
Figure 28 shows the electrical connections between the  
Analyzer and the DUT during a Differential Leakage  
Current test.  
The following outlet conditions apply when performing this  
test:  
Normal Polarity, Closed Earth  
Normal Polarity, Open Earth  
Reversed Polarity, Closed Earth  
Reversed Polarity, Open Earth  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Performing Electrical Safety Tests  
DEVICE UNDER TEST  
MAINS  
ON  
DUT_L1  
DUT_L2  
L1  
APPLIED  
PART  
MD  
MAINS  
L2  
L2  
OPEN  
REV  
POL  
FE  
EARTH  
OPEN  
DUT_PE  
PE  
CONDUCTIVE PART  
TEST  
LEAD  
faw32.eps  
Figure 28. Differential Leakage Current Test Schematic  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
Using the 1210 Adapter  
The 1210 Adapter, an optional accessory, is designed to  
increase the number of lead or applied parts connections  
to the Analyzer from five to 14. The adapter itself ties up to  
ten leads together into a single lead that is plugged into  
one of the input jacks of the Analyzer. The four remaining  
Analyzer input jacks can also be used in conjunction with  
the Adapter.  
Figure 29 shows one application of the Adapter. The  
Defibrillator/Monitor in the example has ten ECG leads,  
two pacer leads, and two defibrillator paddles which need  
to be tested together, and in groups if single function, for  
current leakage per IEC62353. The example shows the  
ECG leads to be snap type connectors and two BJ2ECG  
adapters are shown plugged into the Adapter. If the ECG  
leads did not have snap connectors, then the Universal  
Snap to Banana Adapter can be used to make the  
connections to the Adapter. The common lead from the  
Adapter is plugged into the RA jack (1st jack) of the  
Analyzer. Using four sheathed test leads with alligator  
clips, connect the two defibrillator paddles into the LL and  
LA Analyzer jacks and the two pacer leads into the RL and  
V1 jacks. Using the selection that ties all five Analyzer  
jacks together will test all fourteen leads for leakage  
current. Using the selection of applied part group of 1, 2,  
and 2 allows for testing of groups of applied parts of single  
function.  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Using the 1210 Adapter  
Defibrillator/Pacer  
ESA612  
DEFIB  
OFF  
PACER  
1210  
fis120.eps  
Figure 29. 1210 Adapter Connections  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
When performing an applied parts test using the  
AAMI/NFPA-99 standard, the normal connections of RA,  
LL, LA, and RL are made to their associated Analyzer  
input jacks. Four adapters from the Universal Snap to  
Banana Adapter set will be required for the first four  
connections. The remaining chest leads are connected to  
the Adapter and the common lead from the Adapter is  
plugged into the V1 jack (5th jack) of the Analyzer. See  
Figure. 30. This configuration allows for isolation of the  
RA, LL, LA, and RL leads from each other and the  
remaining chest leads, which are shorted together, while  
performing leakage tests.  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Electrical Safety Analyzer  
Using the 1210 Adapter  
ECG Monitor  
ESA612  
1210  
fis121.eps  
Figure 30. EGC Lead Connection with 1210 Adapter  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
3. Place the probe tips across the unknown voltage and  
read the measurement in the Analyzer’s display.  
Making Point-To-Point Measurements  
The Analyzer can make voltage, resistance, and low  
current measurements through its Point-to-Point function.  
To access the Point-to-Point function menu shown in  
Figure 31, press R. Softkeys F1 through F3 are used  
to select the measurement function.  
The Analyzer will measure up to 300 volts ac.  
Measuring Resistance  
To make a resistance measurement:  
1. Press the softkey labeled Resistance from the  
Point-To-Point menu.  
2. Insert test leads in the RED (V/Ω/A ) and BLACK  
jacks.  
3. Null lead resistance by shorting the leads together  
and pressing the softkey labeled Zero Leads.  
4. Place the probes across the unknown resistance and  
read the measurement in the Analyzer’s display.  
The Analyzer will measure resistances up to 2.0 Ω.  
fis108.jpg  
Figure 31. Point-To-Point Function Menu  
Measuring Voltage  
To make a voltage measurement:  
1. Press the softkey labeled Voltage from the  
Point-To-Point menu.  
2. Insert test leads in the RED (V/Ω/A ) and BLACK  
jacks.  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Simulating ECG Waveforms  
Note  
Measuring Current  
If the ECG monitor/interpreter has banana posts  
instead of snaps, use the optional universal snap  
to banana adapter to connect to the Analyzer.  
The Analyzer can make dc only, ac only, and ac+dc  
current measurements up to 10 mA. To make a current  
measurement:  
To access the ECG Simulation Waveform menu shown in  
Figure 32, press C. From this menu, a number of  
different waveforms are selected through F1, and the rate  
or frequency of the waveform is selected through F2.  
1. Press the softkey labeled Leakage from the  
Point-To-Point menu.  
2. Using G or H select between ac only, dc only, or  
ac+dc measurement mode.  
3. Insert test leads in the RED (V/Ω/A ) and BLACK  
jacks.  
4. Place the leads on the two points the unknown  
current may flow and read the measurement in the  
Analyzer’s display.  
Simulating ECG Waveforms  
The Analyzer is capable of generating various waveforms  
on the applied parts connection posts. These signals are  
used to test the performance characteristics of ECG  
monitors and ECG strip printers. See Figure 33 for proper  
connections between the Analyzer and an ECG monitor.  
For monitors using the snap style connectors, insert the  
BJ2ECG adapter into the connectors at the top of the  
Analyzer and connect the monitor leads to the snap  
connectors on the adapter.  
fis109.jpg  
Figure 32. ECG Waveform Simulation Menu  
To select one of the predefined waveforms, press the  
softkey labeled Wave Form. A scroll box with K next to it  
appears above the softkey label. Use G or H to scroll  
through the different waveforms.  
For all waveforms except VFIB and Triangle, the rate or  
frequency of the waveform is adjusted through the softkey  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
labeled Frequency or Rate. For some waveforms, there  
are more than two frequency or rate selections. For those  
waveforms, pressing the softkey labeled Frequency or  
Rate will open a scroll box above the softkey label with K  
next to it. Use G or H to select the frequency or rate. For  
those waveforms that have only two frequency or rate  
selections, the softkey labeled Frequency or Rate acts as  
a toggle, where each press of the softkey switches to the  
other value.  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Electrical Safety Analyzer  
Simulating ECG Waveforms  
ECG Monitor  
A  
F
L  
A  
RL  
R
R A  
V1  
F
L
N
C
1
L L  
L A  
R L  
C 1  
V 1  
TEST  
BJ2ECG  
Adapter  
fis115.eps  
Figure 33. ECG Monitor Connections  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
5. Press (enter save button) to store the Test Record ID.  
Using Memory  
6. Use G, H, F, and E to enter the date of the Test  
The non-volatile memory of the Analyzer will store up to  
500 measurements or ECG information for each of 100  
different Test Records. Each test record can be recalled to  
the display of the Analyzer or exported to a PC. Data  
Viewer software, available at  
www.flukebiomedical.com/ESA612, is required to upload  
memory data to a PC. Installation and operator  
instructions are also located on the software web page.  
Record.  
7. Press Y to store the date.  
The Test Record ID is displayed in the upper-left  
corner of the display.  
Note  
If no record ID data is entered, a default ID of  
0000000000000001 and a date of --/--/-- is  
assigned to the record.  
Note  
It is recommended that memory data be  
uploaded daily or as often as possible to reduce  
upload time.  
Storing Data into Memory  
Each of the six measurement functions allow data storage  
within the Analyzer. The ECG function allows storage of  
the simulated waveforms only for reference of tests  
performed. To store data, a Test Record ID should be  
created first. To input a new Test Record ID:  
1. Press Q to open the screen shown in Figure 34.  
2. Press G or H to set the first character. Available  
fis114.bmp  
characters are 0-9, A-Z, and space.  
Figure 34. Test Record ID Entry Screen  
3. Press F to move to the next character position.  
8. After selecting a function, push Y to store the  
4. Fill in up to 15 character positions to identify a test  
record.  
measurement or ECG signal under the Test Record  
ID.  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Controlling the Analyzer Remotely  
After performing and saving all the tests for one Test  
Record ID, enter a new Test Record ID to close the  
previous record and start a new one.  
1. Press Q.  
2. Press the softkey labeled More to reveal additional  
menu selections.  
Note  
3. Press the softkey labeled View Memory.  
A previously closed record can not have  
measurements or ECG signals added to it. Data  
can only be stored under the most recently  
opened record.  
4. Select the desired Test Record by pressing G or H  
to scroll through the records listed in the display.  
5. Press the softkey labeled Delete.  
Viewing Memory Data  
Note  
All Test Records can be deleted at this point by  
pressing the softkey labeled Delete All.  
Previously stored data for any Test Record is recalled to  
the screen through the setup menus. To recall data:  
6. When Delete? appears in the display, press the  
softkey labeled Delete to remove the record from  
memory. Otherwise press the softkey labeled Cancel  
to return to the Test Record list and leave the  
selected record in memory.  
1. Press Q.  
2. Press the softkey labeled More to reveal additional  
menu selections.  
3. Press the softkey labeled View Memory.  
4. Select the desired Test Record by pressing G or H  
Note  
to scroll through the records listed in the display.  
Deleting the last or current record does not open  
the previous record for additional data storage.  
Only a new Test Record can be opened for data  
recording after deleting the last record.  
5. Press the softkey labeled View to view the data  
stored for the selected record.  
If the stored data is more than can fit on one screen, press  
the softkey labeled Next Page to view the additional data.  
Controlling the Analyzer Remotely  
Deleting Data from Memory  
Fluke Biomedical Ansur test automation software allows a  
solutions-based approach to complete testing of the  
medical device under test (DUT). Ansur helps create  
standard work using the test template/sequence (which is  
To delete a Test Record and its associated data from  
memory:  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
based on a user written test procedure), and integrates all  
test results into a single test report which can be printed or  
archived. Ansur allows for automatic comparisons to the  
limits of the standard selected, indicating whether results  
are passing or failing. Ansur manages test procedures by  
allowing both manual and visual automated test  
sequences.  
with the options and capabilities needed for the instrument  
being tested.  
Maintenance  
The Analyzer needs little maintenance or special care.  
However, treat it as a calibrated measuring instrument.  
Avoid dropping or other mechanical abuse that could  
cause a shift in the calibrated settings.  
The software works hand-in-hand with Fluke Biomedical  
analyzers and simulators, creating a seamless integration  
for:  
Testing and Replacing the Fuses  
XW Warning  
Visual inspections  
Preventive maintenance  
Work procedures  
Performance tests  
Safety tests  
To prevent electric shock, remove all power  
cords and test leads from the Analyzer before  
opening the fuse door.  
For electrical protection of the equipment outlet, the  
Analyzer uses two fuses, one in the live (L1) line and one  
in the neutral (L2) line. To test the fuses, do the following  
while referring to Figure 35:  
Ansur software utilizes plug-in modules to work with a  
wide array of Fluke Biomedical instruments. The plug-in  
module is a software interface to the Ansur test program.  
The plug-in modules are available for purchase as an  
optional accessory. Plug-ins provide test elements used  
by Ansur. This has the benefit of using the same user  
interface for all analyzers and simulators supported by an  
Ansur plug-in.  
1. Turn the Analyzer so the case bottom is facing up.  
2. Flip up the tilt stand.  
3. Remove the fuse door from the Analyzer by removing  
the screw holding the fuse door with a #2 Phillips  
head screwdriver and lifting the fuse door from the  
Analyzer.  
When a new Fluke Biomedical analyzer or simulator is  
purchased, simply update your existing Ansur software by  
installing a new plug-in. Each plug-in module works only  
4. Remove the fuses from the Analyzer.  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Cleaning the Analyzer  
Cleaning the Analyzer  
XW Warning  
To avoid electric shock, do not clean the  
Analyzer plugged into mains or attached to a  
DUT.  
F1 - F2  
W Caution  
Do not pour fluid onto the Analyzer surface;  
fluid seepage into the electrical circuitry may  
cause the Analyzer to fail.  
W Caution  
Do not use spray cleaners on the Analyzer;  
such action may force cleaning fluid into the  
Analyzer and damage electronic components.  
fis111.eps  
Clean the Analyzer occasionally utilizing a damp cloth and  
mild detergent. Take care to prevent the entrance of  
liquids.  
Figure 35. Fuse Access  
5. Using a multimeter, measure the continuity of each  
fuse.  
Wipe down the adapter cables with the same care. Inspect  
them for damage to and deterioration of the insulation.  
Check the connections for integrity before each use.  
If one or both fuses do not show continuity, replace  
the fuse(s) with fuses that have the same current and  
voltage rating. Appropriate fuse ratings are posted on  
the case bottom label of the Analyzer. Table 6 lists  
available fuses with Fluke Biomedical part numbers.  
6. Reinstall the fuse door and secure it with the screw.  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
Replaceable Parts  
Table 6 lists the replaceable parts for the Analyzer.  
Table 6. Replaceable Parts  
Item  
Fluke Biomedical Part Number  
3334511  
ESA612 Getting Started Manual  
ESA612 Users Manual CD  
3334509  
USA  
2238680  
UK  
2238596  
Australia  
Europe  
2238603  
2238615  
Power Cord  
France/Belgium  
Thailand  
Israel  
2238615  
2238644  
2434122  
Switzerland  
3379149  
Null Post Adapter  
3326842  
Ansur, CD with demo version  
2795488  
5-to-5 Banana jack to ECG (BJ2ECG) adapter  
3359538  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Replaceable Parts  
Table 6. Replaceable Parts (cont.)  
Item  
Fluke Biomedical Part Number  
Carrying Case  
2248650  
1626219  
Data Transfer Cable  
USA  
WT20A 250V Fuse (Time Lag), 1¼x ¼″  
2183691  
109298  
Australia, Switzerland WT10A 250V Fuse (Time Lag), 1¼x ¼″  
Fuse  
Europe, UK, Thailand,  
WT16A 250V Fuse (Time Lag), 6.3 mm x 32 mm  
France/Belgium, Israel  
3321245  
2195732  
15 – 20 A Adapter  
ESA USA/AUS/ISR Accessory Kit:  
Test Lead Set  
3111008  
3111024  
TP1 Test Probe Set  
AC285 Alligator Clip Set  
ESA EUR Accessory Kit:  
Test Lead Set  
TP74 Test Probe Set  
AC285 Alligator Clip Set  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ESA612  
Users Manual  
Accessories  
Table 7 lists the available accessories for the Analyzer.  
Table 7. Accessories  
Item  
Test Leads with Retractable Sheath  
Ground Pin Adapters  
Fluke Biomedical Part Number  
1903307  
2242165  
3392119  
2462072  
1210 - Option  
Universal Snap to Banana Adapter  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Specifications  
Specifications  
Temperature  
Operating ............................................................10 °C to 40 °C (50 °F to 104 °F)  
Storage................................................................-20 °C to 60 °C (-4 °F to 140 °F)  
Humidity.................................................................10 % to 90 % non-condensing  
Altitude  
120 V ac mains supply voltage ...........................5000 m  
230 V ac mains supply voltage ...........................2000 m  
Display....................................................................LCD display  
Communications ...................................................USB device port for computer control  
Modes of Operation...............................................Manual and remote  
Power  
120 Volt power outlet ..........................................90 to 132 V ac rms, 47 to 63 Hz, 20 A maximum  
230 Volt power outlet ..........................................180 to 264 V ac rms, 47 to 63 Hz, 16 A maximum  
Weight ....................................................................1.6 kg (3.5 lb)  
Size .........................................................................28.5 cm x 17.6 cm x 8.4 cm (11.2 in x 6.9 in x 3.3 in)  
Safety Standards  
CE .......................................................................IEC/EN61010-1 2nd Edition; Pollution degree 2  
CSA.....................................................................CAN/CSA-C22.2 No 61010-1; UL61010-1  
Electromagnetic Compatibility Standards (EMC)  
European EMC....................................................EN61326-1  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
ESA612  
Users Manual  
Detailed Specifications  
Test Standard Selections........................................ANSI/AAMI ES-1, IEC62353, IEC60601-1, and AN/NZS 3551  
Voltage  
Ranges (Mains voltage)......................................90.0 to 132.0 V ac rms  
180.0 to 264.0 V ac rms  
Range (Point-to-point voltage)............................0.0 to 300.0 V ac rms  
Accuracy ............................................................. (2 % of reading + 0.2 V)  
Earth Resistance  
Modes .................................................................Two terminal  
Test Current........................................................>200 mA ac  
Range .................................................................0.000 to 2.000 Ω  
Accuracy ............................................................. (2 % of reading + 0.015 Ω)  
Equipment Current  
Range .................................................................0.0 to 20.0 A ac rms  
Accuracy ............................................................. (5 % of reading + (2 counts or 0.2A, whichever is greater))  
Duty cycle ...........................................................15 A to 20 A, 5 min. on/5 min. off  
10 A to 15 A, 7 min. on/3 min. off  
0 A to 10 A, continuous  
Leakage Current  
Modes*................................................................AC+DC (True-rms)  
AC only  
DC only  
* Modes: AC+DC, AC only, and DC only available for all leakages with exception of MAP  
that are available in True-rms (shown as AC+DC)  
Patient Load Selection........................................AAMI ES1-1993 Fig. 1  
IEC 60601: Fig. 15  
Crest factor .........................................................3  
62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Electrical Safety Analyzer  
Detailed Specifications  
Ranges................................................................0.0 to 199.9 μA  
200 to 1999 μA  
2.00 to 10.00 mA  
Accuracy  
DC to 1 kHz..................................................... (1 % of reading + (1 μA or 1 LSD, whichever is greater))  
1 to 100 kHz.................................................... (2 % of reading + (1 μA or 1 LSD, whichever is greater))  
100 kHz to 1 MHz............................................ (5 % of reading + (1 μA or 1 LSD, whichever is greater))  
Mains on applied part test voltage ......................100 % 5 % of Mains for AAMI, current limited to 1 mA 25 % per AAMI  
100 % 5 % of Mains for 62353 current limited to 3.5 mA 25 % per 62353  
100 % 5 % of Mains for 60601 current limited to 7.5 mA 25 % per 60601  
Differential leakage  
Ranges................................................................10 to 199 μA  
200 to 2000 μA  
2.00 to 20.00 mA  
Accuracy.............................................................. (10 % of reading + (2 counts or 20 μA, whichever is greater))  
Insulation resistance  
Ranges................................................................0.5 to 20.0 MΩ  
20.0 to 100.0 MΩ  
Accuracy  
20 MΩ Range.................................................. (2 % of reading + 0.2 MΩ)  
100 MΩ Range................................................ (7.5 % of reading + 0.2 MΩ)  
Source test voltage .............................................500 V dc (+20 %, -0 %) 1.5 0.25 mA short-circuit current or 250 V dc selectable  
Maximum load capacitance.................................1 μF  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ESA612  
Users Manual  
ECG Performance Waveforms  
Accuracy  
Frequency....................................................... 2 %  
Amplitude........................................................ 5 % of 2 Hz square wave only, fixed @ 1 mV Lead II configuration  
Waveforms  
ECG Complex.................................................30, 60, 120, 180, and 240 BPM  
Ventricular Fibrillation  
Square wave (50 % duty cycle) ......................0.125 Hz and 2 Hz  
Sine wave .......................................................10, 40, 50, 60, and 100 Hz  
Triangle wave..................................................2 Hz  
Pulse (63 ms pulse width)...............................30 BPM and 60 BPM  
64  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Euro Pro Vacuum Cleaner V1055H User Manual
Extron electronic Projector Accessories 70 526 03 User Manual
Faber Ventilation Hood 630006369 User Manual
Fisher Price Baby Toy V2770 User Manual
Fisher Price Mobility Aid H6372 User Manual
Fisher Price Riding Toy W1442 User Manual
FujiFilm Camcorder X Pro1 User Manual
Fusionbrands GPS Receiver MS BT100 User Manual
GE Answering Machine 29861 User Manual
Gefen Switch 2x1 DVI KVM Switcher User Manual