Fluke Stereo Amplifier 943 227 15 User Manual

®
Victoreen  
Digital Preamplifier, Model 943-227-15,for use with  
Model 943-27 Current Mode Beta Detector  
Operator Manual  
February 2008  
Manual No. 943-227-15-1, Rev. 3  
©2007, 2008 Fluke Corporation, All rights reserved. Printed in U.S.A.  
All product names are trademarks of their respective companies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
Section 1:  
1.1  
General Information............................................................................................................. 1-1  
General Description ............................................................................................................... 1-1  
Application.............................................................................................................................. 1-1  
Specifications......................................................................................................................... 1-2  
Auxiliary Equipment ............................................................................................................... 1-6  
Recommend Spare Parts....................................................................................................... 1-6  
Receiving Inspection.............................................................................................................. 1-6  
Storage................................................................................................................................... 1-7  
1.2  
1.3  
1.4  
1.5  
1.6  
1.7  
Section 2:  
2.1  
Theory of Operation............................................................................................................. 2-1  
Functional Description............................................................................................................ 2-1  
Main Circuit Boards................................................................................................................ 2-1  
977-200-15M Electrometer Board.......................................................................................... 2-1  
977-210-10M Interface Board................................................................................................ 2-4  
Preamplifier Switch & Jumper Settings.................................................................................. 2-5  
2.2  
2.3  
2.4  
2.5  
Section 3:  
3.1  
Operation .............................................................................................................................. 3-1  
Installation.............................................................................................................................. 3-1  
Set-Up.................................................................................................................................... 3-2  
Operation ............................................................................................................................... 3-3  
Configuration Functions......................................................................................................... 3-4  
Set point Error Codes............................................................................................................. 3-5  
Software Calculations ............................................................................................................ 3-5  
3.2  
3.3  
3.4  
3.5  
3.6  
Section 4:  
4.1  
Maintenance, Calibration and Troubleshooting ............................................................... 4-1  
Maintenance........................................................................................................................... 4-1  
Calibration.............................................................................................................................. 4-1  
Troubleshooting ..................................................................................................................... 4-2  
4.2  
4.3  
Appendix A:  
A.1  
Connector Designations .....................................................................................................A-1  
Connector Designations.........................................................................................................A-1  
Appendix B:  
B.1  
Applicable Drawings............................................................................................................B-1  
Applicable Drawings...............................................................................................................B-1  
Appendix C:  
C.1  
Bill of Materials.....................................................................................................................C-1  
Bill of Materials.......................................................................................................................C-1  
Appendix D:  
D.1  
Cable Termination Instructions..........................................................................................D-1  
Cable Termination Instructions ..............................................................................................D-1  
Appendix E:  
E.1  
Modification Sheets/Engineering Instructions............................ .....................................E-1  
Modification Sheets/Engineering Instructions........................................................................E-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
(Blank page)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General Information  
General Description  
1
Section 1  
General Information  
1.1 General Description  
The Victoreen® Model 943-227-15 Digital Preamplifier, is designed to operate with the Model 943-27,  
Current Mode Beta Scintillation detector and the Model 942A-200C Series Digital Ratemeter (UDR) or the  
Model 960 Series Digital Process Control system. When connected to the beta sensitive scintillation  
detector, it comprises a monitoring system which operates in the range of 10-3 to 105 uCi/cc (or 1E0 to 1E8  
Pico amperes). The UDR provides display, control, and annunciator functions for the monitoring system.  
The preamplifier contains a microprocessor controlled, auto-zeroing, integrating electrometer, a  
programmable gain amplifier, an analog-to-digital converter and an asynchronous serial communications  
interface. Communications with the UDR is accomplished via the optically isolated Victoreen Serial  
Communication Loop driver/receiver circuitry.  
Interconnection of the detector and preamplifier is accomplished utilizing two, six (6) foot coaxial cables.  
One cable supplies the high voltage for the detector (negative polarity). The other cable is used for the  
detector output signal. Interconnection between the preamplifier and the UDR or 960 system is  
accomplished via a six-conductor cable and a high voltage coaxial cable.  
The detector operates at a potential of approximately (-) 500 VDC, and produces a nominal DC output  
current of 1 x 10-12 amperes to 1 x 10-4 amperes. Depending on sampling geometry and detector set up,  
this corresponds to a range of 1 x 10-3 to 1 x 10+5 uCi/cc.  
The preamplifier is a self-contained electronics unit that processes the DC input signal from the detector,  
digitizes the signal, and transmits the digitized value to the readout via a modified RS232C compatible  
serial data communication port. It contains two printed circuit boards (mounted on the rear of the  
enclosure cover, a terminal block for terminating the field cable from its associated controller and two  
connectors for terminating the detector Signal and HV coaxial cables. The larger of the two printed circuit  
boards is the 977-210-10M Serial Interface board. This board controls the communications with the  
controller. The smaller board is the 977-200-15M Electrometer. It contains the microprocessor and  
associated circuitry for converting the detector output current into a digital value.  
For additional information on the detector or readout device, refer to the instruction manual for the Model  
943-27-31 detector, the Model 942A-200C UDR, or the applicable Model 960 System manual.  
1.2 Application  
The Model 943-227-15 Digital Preamplifier is used with the Victoreen 942A-200C or Model 960  
microprocessor based radiation monitors. The Model 942A-200C or 960 system provides the +15 VDC  
preamplifier electronics power, (-) 500 VDC detector power and the RS232C serial communications  
support. The digital preamplifier is available in two serial port configurations. For applications less than  
50 feet, the Model 943-227-15, configured for RS232C communications, is used. For applications  
requiring signal transmission distances of up to 1 mile, the Model 943-227-15VL, configured for Victoreen  
Loop communications is used. Table 1-1 is a list of available configurations and compatible 9XX series  
equipment. Please contact Fluke Biomedical Radiation Management Services for additional information.  
1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
Table 1-1. 943-227-15 Variations and Compatible Equipment  
Preamplifier  
Variations  
Application  
Detector  
Readout  
943-227-15 Digital  
Preamplifier  
Transmission  
Distances  
Less than 50 feet  
943-27-31 Beta  
Detector,  
Current Mode  
Operation  
942A-200C Digital Ratemeter  
or Model 960 Digital Process  
Control System  
943-227-15VL Digital  
Preamplifier  
Transmission  
Distances up to 1  
mile  
943-27-31 Beta  
Detector,  
Current Mode  
942A-200C Digital Ratemeter  
or Model 960 Digital Process  
Control System  
Operation  
943-227-15VL-M1  
Digital Preamplifier  
Transmission  
Distances up to 1  
mile; 2.0 inch conduit Current Mode  
field cable interface Operation  
943-27-31 Beta  
Detector,  
942A-200C Digital Ratemeter  
or Model 960 Digital Process  
Control System  
1.3 Specifications  
General specifications for the Model 943-227-15 Digital Preamplifier are listed below. The 943-227-15  
series products are designed for nuclear applications, and any repairs to it by personnel not qualified to  
ANSI 45.2.6 1978, Skill Level II may void the nuclear rating.  
If a problem develops, the preamplifier may either be returned to the factory for service, or repaired by a  
qualified technician.  
1-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General Information  
1
Application  
Table 1-2. General Specifications for the 943-227-15 Digital Preamplifier  
Feature:  
Specification:  
+ 15 v @ 20 mA  
Preamplifier Power  
DC current, 1E-12 to 1E-3 amperes from 943-27 Current Mode Beta  
Scintillation detector  
Preamplifier Input  
943-227-15, RS232C;  
943-227-15VL, Modified RS232C serial data for transmission distances up  
to 5,280 feet.  
± 10% of the input current  
Preamplifier Output  
Electronic Accuracy  
Decade / Time:  
1 / 80 sec  
2 / 32 sec  
3 / 16 sec  
4 / 4 sec  
Response Time  
5 / 2 sec  
6 / 0.9 sec  
7 / 0.2 sec  
8 / 0.2 sec  
943-227-15 or 943-227-15VL:  
8.0 in. x10.0 in. x 4.0 in.  
(20.3 cm x 25.4 cm x 10.2 cm)  
Dimensions  
(H x W x D)  
943-227-15-M1 or 943-227-15VL-M1:  
8.0 in. x10.0 in. x 6.0 in.  
(20.3 cm x 25.4 cm x 15.2 cm)  
Approximately 8 lbs. (3.6 kg)  
Weight  
Gasketed, NEMA 4, Steel enclosure with hinged cover  
Housing  
32 °F to 122 °F  
0 °C to + 50 °C  
Operating Temperature  
0 to 99 % non-condensing  
Relative Humidity  
Victoreen 943-27 Current Mode Beta scintillation detector only  
Compatible Detectors  
Detector cables enter thru “Seal Grip” fittings; Mating SHV receptacle  
provided for HV; Mating BNC receptacle provided for Signal  
Victoreen Loop Communications: 6 conductors, 18 AWG (2, TwPr, 2 Single  
conductor) with overall shield. For max transmission distance, 16 AWG is  
recommended;  
Detector Interface  
Cable Requirements  
RS232C communications: 8 Conductors, 18 AWG (4 TwPr) with overall  
shield.  
For the Model 943-227-15VL-M1, refer to Modification description provided  
in Appendix E  
Modification  
Information  
1-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
Table 1-3. Preamplifier Serial Interface Board 977-210-10M Switch Settings  
Switch Normal Switch Function  
Op.  
SW1  
Open Microprocessor Reset switch, momentary contact  
SW2  
N/A  
Not supplied  
SW3-1  
SW3-2  
SW3-3  
ON  
OFF  
OFF  
OFF Selects 300 baud 1/ON Selects 4800 baud (normal setting)  
OFF • (No function is implemented in the 977-210-10M)  
OFF Selects Rate Mode (normal) • (Selects Maintenance Mode for  
factory PGA and Auto Zero set-up only)  
SW3-4  
OFF  
OFF • (No function is implemented in the 977-210-10M)  
NOTE  
SW3 is a 4-position DIP switch on the 977-210-10M serial  
interface board.  
Table 1-4. Jumper Positions for Preamplifier Serial Interface Board 977-210-10M  
Jumper  
Normal Setting Function  
Alternate Setting Function  
JMP1  
A to B  
Enables Victoreen Loop B to C  
Enables RS232C  
(normal)  
JMP2  
N/A  
Deleted, Not Used  
N/A  
Deleted, Not Used  
Table 1-5. Jumper Positions for Preamplifier Electrometer Board 977-200-15M  
Jumper  
JMP1  
Normal Setting Function  
A to B Normal Operation  
Cal. Position Function  
B to C PGA Offset Adjust *  
* Programmable Gain Amplifier requires a special test PROM and is a factory adjustment  
1-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General Information  
1
Specifications  
Table 1-6. Field Electrical Connections, 943-227-15 and 943-227-15VL  
943-227-15  
Description  
943-27-15VL  
Description  
Connects To  
977-210-10M  
Connects To  
977-210-10M  
TB1-  
+ VL, 15  
VDC  
J4-1  
J3-1  
1
XMIT  
J3-2  
J3-3  
J3-4  
J3-5  
J3-6  
2
3
DTR  
+ TX  
J4-2  
J4-3  
GND  
- TX  
4
REC  
- VL  
J4-4  
Not Used  
Not Used  
+ 15 VDC  
Ground  
- 15 VDC  
Not Used  
Not Used  
5
+ RX  
Not Used  
J3-7  
6
- RX  
7
Not Used  
Ground  
Not Used  
Not Used  
Not Used  
J3-8  
J3-8  
8
9
J3-4  
Not Used  
J3-11  
10  
Not Used  
Detector  
Connections  
Detector  
Connections  
Detector  
From Field  
coaxial cable  
From Field  
coaxial cable  
HV  
Detector MHV  
MHV  
Factory  
Factory  
Terminated,  
To 977-200-  
15M, R41  
Detector  
BNC  
Terminated,  
To 977-200-  
15M, R41  
SIG  
Detector BNC  
Table 1-7. 977-210-10M LEDs  
LED #  
Indicates  
LED1  
+15 VDC is being supplied to the circuit board  
from the UDR when ON  
LED2  
Communications – Receive;  
ON (bright): both + 15 and -15 volt loops  
present  
ON (dim): Only one loop supply present  
OFF: no loop voltage from either supply  
Table 1-8. 977-210-10M Test points  
Test Point  
Function  
TP2  
RS232 Transmit  
1-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
1.4 Auxiliary Equipment  
Model  
Description  
None  
1.5 Recommended Spare Parts  
Table 1-9. Recommended Spare Parts List for the Model 943-227-15 Digital Preamplifier  
Part Number Description  
Used On  
977-200-15M Electrometer Circuit Board  
977-210-10M Serial Interface Circuit Board  
-
-
92-7072  
450-1-0531  
Fuse, F1, F2, 1Amp, Communication Fuse  
Dryer, Air (5 Grams)  
977-210-10M  
977-200-15  
1.6 Receiving Inspection  
Upon receipt of the unit:  
1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier and  
notify the Fluke Biomedical Radiation Management Services Customer Service Department.  
Fluke Biomedical  
Radiation Management Service  
6045 Cochran Road  
Cleveland, Ohio 44139  
Phone: 440.248.9300  
Fax: 440.542.3682  
2. Remove the contents from the packing material.  
3. Verify that all items listed on the packing list have been received and are in good condition.  
NOTE  
If any of the listed items are missing or damaged, notify the  
Fluke Biomedical Radiation Management Service Customer  
Service Department.  
1-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General Information  
Spare Parts/Receiving Inspection  
1
1.7 Storage  
Storage of Fluke Biomedical instruments must comply with Level B storage requirements as outlined in  
ANSI N45.2.2 (1972) Section 6.1.2(.2). The storage area shall comply with ANSI N45.2.2 (1972) Section  
6.2 Storage Area, Paragraphs 6.2.1 through 6.2.5. Housekeeping shall conform to ANSI N45.2.3 (1972).  
Level B components shall be stored within a fire resistant, tear resistant, weather tight enclosure, in a  
well-ventilated building or equivalent.  
Storage of Fluke Biomedical instruments must comply with the following:  
1. Inspection and examination of items in storage must be in accordance with ANSI N45.2.2 (1972)  
Section 6.4.1.  
2. Requirements for proper storage must be documented and written procedures or instructions must  
be established.  
3. In the event of fire, post-fire evaluation must be in accordance with ANSI N45.2.2 (1972), Section  
6.4.3.  
4. Removal of items from storage must be in accordance with ANSI N45.2.2 (1972), Sections 6.5 and  
6.6.  
1-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
(Blank Page)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Theory of Operation  
Functional Description  
2
Section 2  
Theory of Operation  
2.1 Functional Description  
The Model 943-227-15 Digital Preamplifier is an eight-decade, linear reading radiation. The preamplifier  
is sensitive to moisture and is mounted in a NEMA 4 sealed enclosure. The preamplifier box contains a  
desiccant package to remove moisture from the interior of the preamplifier. The desiccant has a color  
indicator to show when it should be replaced when the color changes to blue.  
The detector electrometer converts the current from the detector to a voltage which is measured by the  
analog to digital converter of the preamplifier. Under program control, this measurement is made every  
50 milliseconds on the first five decades and every 100 milliseconds on the upper three decades.  
The electrometer is auto-ranging and auto-zeroing. An internal coarse zero control (factory adjusted) is  
provided. The microprocessor in the preamplifier performs data collection, integration and multiplication  
by a stored calibration factor, range changing and communication with the Model 942A-200C Digital  
Ratemeter or Model 960 Digital Process Radiation monitoring system.  
2.2 Main Circuit Boards  
The wide range ion chamber detector block diagram is shown in figure 2-1. Detector electronics are  
contained within two circuit board assemblies located in the preamplifier enclosure: the 977-210-10 High  
Voltage/Communications Interface circuit board assembly and the 977-200-15 Electrometer circuit board  
assembly. Schematic diagrams are located in Appendix B.  
The 977-200-15M Electrometer circuit board contains the microprocessor, the EPROM, the integrating  
electrometer, the programmable gain amplifier, the analog to digital (A/D) converter, and the auto zero  
DAC circuit.  
The 977-210-10M Communications Interface circuit board contains the dc voltage regulator, the BDC/DC  
converter, the asynchronous communication interface adapter (ACIA), the Victoreen communication loop  
driver/receiver, the RS232 communications driver/receiver, and the communication loop fuses.  
2.3 977-200-15M Electrometer  
Refer to the electrometer schematic 977-200-18M located in Appendix B. Z7 is a Texas Instruments TMS  
70C00 microprocessor, which is utilized to control detector operation. Z8 latches port C of the  
microprocessor to form the lower order bits of memory address. Z10 is a 27C64 EPROM containing the  
operational firmware for the detector. Z18 is an optional RAM socket which is not utilized for this model  
detector. Z9 is a decoder for the upper order address bits. Z11 is a latch used to store control bits to  
select the gain of the programmable gain amplifier formed by the combination of Z5, Z4, and RN2.  
Available gains are 1 through 128 in binary increments 1, 2, 4, . . . 128. Z3 is an eight-bit analog to digital  
converter used to digitize the output of the programmable gain amplifier. Analog switch Z17 pin 11  
selects either the integrating electrometer or the high voltage sense line as the input to the programmable  
gain amplifier is not active in this application. Refer to Table 2-1 for preamplifier register addresses and  
Table 2-2 for electrometer control bit assignments.  
2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
0/-12 Vdc  
0 - 10 Vdc  
DAC  
ALARM  
OUTPUT  
INTEGRATING  
ELECTROMETER  
EPROM  
MICROPROCESSOR  
ACIA  
PGA  
ANALOG  
SWITCH  
ADC  
SIGNAL  
CONTROL  
HIGH VOLTAGE  
POWER SUPPLY  
HIGH VOLTAGE  
Resistor  
+VL  
+TX  
DETECTOR  
ASSEMBLY  
VICO  
LOOP  
RS232  
Resistor  
-TX  
+RX  
-RX  
-VL  
MICROPROCESSOR  
CONTROL  
SIGNAL  
Figure 2-1. Typical Detector Electronics Block Diagram  
Table 2-1. Preamplifier Register Addresses  
Function  
Hex Address  
2000  
4000  
6000  
A000  
Electrometer Control  
ADC Convert CMD  
ADC Read Data  
ACIA T/R Data  
ACIA Status Register  
ACIA Command Register  
ACIA Control Register  
Input SW3  
A001  
A002  
A003  
A008  
RAM (not used)  
EPROM  
C000-DFFF  
E000-FFFF  
2-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Theory of Operation  
Main Circuit Boards  
2
Table 2-2. Electrometer Control Bit Assignments  
Bit Position  
7
6
5
4
3
2
0
0
0
0
1
1
1
1
X
X
X
X
1
0
0
1
1
0
0
1
1
X
X
X
X
0
0
1
0
1
0
1
0
1
X
X
X
X
Description  
Gain = 1  
Gain = 2  
Gain = 4  
Gain = 8  
Gain = 16  
Gain = 32  
Gain = 64  
Gain = 128  
Unshort capacitor  
Short capacitor  
Digitize electrometer  
Digitize high voltage  
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
0
X
X
X
X
X
X
X
X
0
1
X
X
1
X = Bit position does not matter (1 or 0)  
The integrating electrometer is formed by the combination of current source Z6, dual MOSFET Q4,  
operational amplifier Z12, transistor Q3, and analog switch Z17, pin 15.  
Counter Z15 and ladder network RN3 form the auto-zero portion of the integrating electrometer. Refer to  
section 4 for auto-zero adjustment instructions (note this is a factory only adjustment).  
Dual MOSFET Q4 acts as a low leakage input buffer to the electrometer. C17 is an integrating capacitor,  
while Z17, pin 15 and Q3 are the shorting elements of the integrator.  
There is a hysteresis built into the range changing function of the electrometer. The range change-up  
occurs at the decade points whereas the range change-down occurs at 80% of the decade points. This  
hysteresis results in the display of only two significant digits when in the upper 20% of a decade.  
Whether or not this occurs depends on which direction the radiation level is trending. An increasing trend  
will result in three significant digits while a decreasing trend will result in two significant digits.  
The response times for a change in reading within the same decade are listed in Table 2-1. If a change  
of 100 to 1000 times the current reading occurs, the instrument will range change within a single 50-  
millisecond clock cycle.  
Table 2-3. Response Time  
DECADE  
RANGE  
RESPONSE TIME  
80 sec.  
32 sec.  
16 sec.  
4 sec.  
1
2
3
4
5
6
7
8
0.1 to 1.0 mR/h  
1.0 to 10 mR/h  
10 to 100 mR/h  
100 to 1000 mR/h  
1.0 to 10 R/h  
10 to 100 R/h  
100 to 1000 R/h  
1.0 k to 10 kR/h  
2 sec.  
0.9 sec.  
0.2 sec.  
0.2 sec.  
2-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
The following Test are provided on the 977-200-15M board:  
Table 2-4. 977-200-15M Test Points  
Test Point  
TP1  
Function  
Electrometer Output  
TP2  
Analog +5 VDC supply  
TP3  
- 5 VDC supply  
TP4  
TP5  
TP6  
TP7  
Digital +5 VDC supply (From 977-210-10M)  
High Voltage Sense (Not Used)  
+ 12 VDC supply (From 977-210-10M)  
DC Ground  
TP8  
TP9  
TP10  
Programmable Gain Amplifier Output  
Microprocessor Clock Signal  
Chip Enable - EPROM  
2.4 977-210-10M Interface Board  
Refer to the Communications Interface, P/N 977-210-13M, schematic diagram in Appendix B. U12 is an  
asynchronous communications interface adapter (ACIA) which communicates with the UDR or 960  
system. U1 is an analog switch used to select either the Victoreen loop or the RS232 driver/receiver for  
external communications. Optical isolation U10 isolates data transmitted on the Victoreen loop while the  
circuitry comprised of Q6, Q7, Q14, and Q18 are the actual loop drivers. Optical isolator U8 isolates the  
receive data from the Victoreen loop.  
U2 and U3 are, respectively, the receiver and driver circuits for the RS232 port. U14 decodes address  
block A000 to provide chip enables to the ACIA switch input register U6.  
Regulated power for the electronics (+/- 12 VDC) is derived from three terminal regulators VR3 and VR4.  
The ±5 Vdc logic power is derived from the three terminal regulators VR1 and VR2.  
Power for the detector is normally provided by the electronic high voltage power supply in the associated  
Model 942A-200C UDR or Model 960 System.  
The following LEDs are provided to indicate operation of certain functions of the 977-210-10M board:  
Table 2-5. 977-210-10M LEDs  
LED #  
Indicates  
LED1  
+15 VDC is being supplied to the circuit board  
from the UDR when ON  
LED2  
Communications – Receive;  
ON (bright): both + 15 and -15 volt loops  
present  
ON (dim): Only one loop supply present  
OFF: no loop voltage from either supply  
The following Test are provided on the 977-210-10M board:  
Table 2-6. 977-210-10M Test points  
Test Point  
Function  
TP2  
RS232 Transmit  
2-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Theory of Operation  
2
Communications  
Victoreen Communications Loop – 943-227-15VL  
Communication between the detector preamplifier and the UDR or 960 controller may be performed via  
Victoreen loop, a serial asynchronous communications interface. The communication utilizes Victoreen  
protocol and is intended for use in electrically hostile environments where high noise immunity is required.  
Noise immunity is achieved using differential line driving, and electrically isolating the end of each  
transmit/receive pair with opto-isolators. Using a ±15 Vdc communications power supply, each 0 to 1  
transition is equivalent to a 30-volt swing in loop voltage. Refer to the Model 942-200-80-1  
Communications Board instruction manual for further information.  
Jumper JP1 controls the communication communications method employed. JP1, A-B selects Victoreen  
loop communications.  
RS232C Communications – 943-227-15  
Communication between the detector preamplifier and the UDR or 960 controller may be performed via  
RS232C serial asynchronous communications interface. Jumper JP1 controls the communication  
communications method employed.  
JP1, B-C selects RS232C communications.  
2.5 Preamplifier Switch & Jumper Settings  
Switch and jumper settings for the preamplifier are summarized in Table 2-7, 2-8, and 2-9.  
Table 2-7. Preamplifier Serial Interface Board 977-210-10M Switch Settings  
Switch Normal Switch Function  
Op.  
SW1  
Open Microprocessor Reset switch, momentary contact  
SW2  
N/A  
Not supplied  
SW3-1  
SW3-2  
SW3-3  
ON  
OFF  
OFF  
OFF Selects 300 baud 1/ON Selects 4800 baud (normal setting)  
OFF • (No function is implemented in the 977-210-10M)  
OFF Selects Rate Mode (normal) • (Selects Maintenance Mode for  
factory PGA and Auto Zero set-up only)  
SW3-4  
OFF  
OFF • (No function is implemented in the 977-210-10M)  
NOTE  
SW3 is a 4-position DIP switch on the 977-210-10M serial  
interface board.  
2-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
Table 2-8. Jumper Positions for Preamplifier Serial Interface Board 977-210-10M  
Jumper  
Normal  
Setting  
A to B  
Function  
Alternate Function  
Setting  
JMP1  
B to C  
Enables RS232C  
Enables Victoreen Loop  
(normal)  
JMP2  
N/A  
Deleted, Not Used  
N/A  
Deleted, Not Used  
Table 2-9. Jumper Positions for Preamplifier Electrometer Board 977-200-15M  
Jumper  
Normal  
Setting  
A to B  
Function  
Cal.  
Position  
B to C  
Function  
JMP1  
Normal Operation  
PGA Offset Adjust *  
* Programmable Gain Amplifier requires a special test PROM and is a factory adjustment  
2-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Installation  
3
Section 3  
Operation  
3.1 Installation  
Installation consists of mounting the equipment, making the required electrical connections, and entering  
the desired set points.  
CAUTION  
Remove all power prior to installing the  
Preamplifier.  
Preamplifier Mounting  
The preamplifier is a self-contained unit that is designed for mounting on a wall, or structurally sound  
surface, within 6 feet of the detector. The preamplifier must be located in an area that is accessible for  
future maintenance.  
CAUTION  
Remove all power prior to connecting field wiring.  
Electrical Interface  
Electrical connections at the preamplifier include connecting the detector high voltage and signal coaxial  
cables and the field cable from the associated controller.  
The detector high voltage and signal cables enter the preamplifier through “Seal Grip” type penetrations  
on the enclosure body. The “Seal Grips” contain a rubber gland that is intended to grip the coaxial cables  
to prevent moisture entry into the enclosure. The “Seal Grips” are an integral part of the detector cables.  
The “Seal Grip” enclosure nut is removed from the fitting. The insulating washer stay with the fitting. The  
detector signal and high voltage cables, with connectors and “Seal Grip” fitting nuts are then inserted into  
the holes preamplifier enclosure. The “Seal Grip: end fittings are then re-installed securely against the  
wall of the enclosure.  
The detector Signal (BNC) and High Voltage (SHV) cables are then terminated on the bulkhead  
connectors located on the enclosure inner panel.  
On the 943-227-15 or 943-227-15VL, the field cable is terminated on a “MS” type connector. A mating  
“MS” style connector supplied with the detector. It is not necessary to open the preamplifier enclosure to  
complete the interconnection to the associated controller.  
On the 943-227-15VL-M1, a 2.0-inch conduit hub is provided to interface with the field conduit system  
and cable. The field conduit is connected to the preamplifier conduit hub and the field cable is routed  
through the conduit and into the preamplifier enclosure. The field signal cable is then terminated on the  
terminal block in the enclosure. Refer to the applicable wiring diagram or Table 3-1. The high voltage  
coax cable is then terminated on the mating SHV connector. Refer to Appendix D for further information.  
A desiccant bag is provided in the enclosure to absorb moisture. The desiccant bags should be checked  
for moisture absorption prior to placing into service. Each bag contains a color indicator stripe that turns  
red when it should be replaced.  
3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
NOTE  
The electrometer circuitry in the preamplifier is highly sensitive to  
moisture and physical damage. Use extreme care when the  
preamplifier case is opened as the high impedance electrometer  
circuitry is easily damaged.  
Do not leave the preamplifier  
enclosure in the open position for any extended period of time.  
The preamplifier terminations are listed below.  
Table 3-1. 943-227-15 Preamplifier Electrical Connections  
943-227-15  
943-27-15VL  
Connects To  
977-210-10M  
Connects To  
977-210-10M  
TB1-  
Description  
Description  
+ VL, 15  
VDC  
J4-1  
J3-1  
1
XMIT  
J3-2  
J3-3  
J3-4  
J3-5  
J3-6  
2
3
DTR  
+ TX  
J4-2  
J4-3  
GND  
- TX  
4
REC  
- VL  
J4-4  
Not Used  
Not Used  
+ 15 VDC  
Ground  
- 15 VDC  
Not Used  
Not Used  
5
+ RX  
Not Used  
J3-7  
6
- RX  
7
Not Used  
Ground  
Not Used  
Not Used  
Not Used  
J3-8  
J3-8  
8
9
J3-4  
Not Used  
J3-11  
10  
Not Used  
Detector  
Connections  
Detector  
Connections  
Detector  
From Field  
coaxial cable  
From Field  
coaxial cable  
HV  
Detector MHV  
MHV  
Factory  
Factory  
Terminated,  
To 977-200-  
15M, R41  
Detector  
BNC  
Terminated,  
To 977-200-  
15M, R41  
SIG  
Detector BNC  
3.2 Set-up  
To place the system in operation, the following steps should be performed:  
NOTE  
Ensure you have read and fully understand  
Section 1, 2 and 3 prior to continuing.  
1. Verify that jumpers and DIP switches and jumpers on the preamplifier, are set for the operational  
features desired. Refer to Section 1, Table 1-4.  
2. Refer to the applicable controller manual prior to powering up the controller.  
3-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
3-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Configuration  
3
3.3 Operation  
Operation of the 943-227-15 Digital Preamplifier is automatic and is controlled by the Model 942A-200C  
Digital Ratemeter or Model 960 Digital Process Radiation monitoring system. Once the power to the  
controlling device is turned on, the preamplifier begins operation. No operator interaction is required.  
Refer to the instruction manual for the applicable controller, detector and the Model 942-200-80  
Communications Option Board for further information.  
Normal Operation  
If the measured radiation field is within the range of the detector during power-up, the uCi/cc value will be  
displayed on the controller. The digital display will update once per second.  
Alarms  
The radiation alarms are initiated by the applicable controller. Refer to the applicable controller manual.  
Fail Alarm  
There are two equipment failure conditions that can produce a FAIL output or an error display on the  
associated controller. The fail condition is true whenever any equipment failure is detected and false  
when no equipment failures are detected.  
The following are the fail alarms provided by the 943-227-15 Preamplifier:  
1. POWER Failure  
2. Loop Test Failure  
To return the channel to normal operation after a FAIL alarm, the condition, which caused the alarm, must  
be located and corrected.  
Power Failure  
If the power to the controller is lost, the bar graph, alarm indicators, and the display are blanked (turned  
off). The HIGH, WARN, and FAIL relay coils will de-energize. If power to the preamplifier only is lost, or if  
there is a break in the power cable, operation of the microprocessor will stop. The display on the  
controller will then cease to update. If power to the controller is not lost, a break in the interconnecting  
wiring or the internal communication power fuses on the 977-210-10M Interface board should be  
checked. Loss of power to the preamplifier will also result in a loop communication failure.  
Loop Failure  
If the UDR or 960 controller does not receive a valid message from the preamplifier in ten seconds, a loop  
failure is detected and an E0009 code will normally be displayed. Note the error code displayed is  
defined by the controlling device, and not by the preamplifier.  
The loop failure can originate from bad seating of the serial communications board in the UDR or 960CD,  
bad connections in the preamplifier, bad connections at the P2 connector of the UDR, a baud rate  
mismatch between the setting on the serial board and that in the preamplifier, an invalid address setting  
on the serial board, or blown communication loop fuses.  
The baud rate of the preamplifier may be set to either 300 or 4800 baud. SW3-1 on the 977-interface  
board must be in the ON position to select a baud rate of 4800 baud, the normal setting.  
3-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
For UDR systems, the address of the 942-200-80 Communications board in the UDR must be set 4040  
and the baud rate must be set to 4800 baud. If the wiring of the Communications option board to the P2  
connector is suspected, a loop test can be performed as described in "Loop Test Mode". The instruction  
manual for the Communications option board (P/N 942-200-80-1) contains information on how to set up  
the serial board in the UDR.  
Firmware Version  
This system requires a PROM for operation. For the part number and the latest revision of the PROM,  
see the Factory Test Data Sheet. This Manual has been prepared for use with the P/N 227LNK, Rev. 2,  
PROM.  
3.4 Configuration Functions  
The preamplifier configuration jumpers and required settings for this application are listed below:  
Table 3-2. Preamplifier Serial Interface Board 977-210-10M Switch Settings  
Switch Normal Switch Function  
Op.  
SW1  
Open Microprocessor Reset switch, momentary contact  
SW2  
N/A  
Not supplied  
SW3-1  
SW3-2  
SW3-3  
ON  
OFF  
OFF  
OFF Selects 300 baud • 1/ON Selects 4800 baud (normal setting)  
OFF • (No function is implemented in the 977-210-10M)  
OFF Selects Rate Mode (normal) • (Selects Maintenance Mode for  
factory PGA and Auto Zero set-up only)  
SW3-4  
OFF  
OFF • (No function is implemented in the 977-210-10M)  
NOTE  
SW3 is a 4-position DIP switch on the 977-210-10M serial  
interface board.  
Table 3-3. Jumper Positions for Preamplifier Serial Interface Board 977-210-10M  
Jumper  
Normal  
Setting  
A to B  
Function  
Alternate Function  
Setting  
JMP1  
B to C  
Enables RS232C  
Enables Victoreen Loop  
(normal)  
JMP2  
N/A  
Deleted, Not Used  
N/A  
Deleted, Not Used  
Table 3-4. Jumper Positions for Preamplifier Electrometer Board 977-200-15M  
Jumper  
Normal  
Setting  
A to B  
Function  
Cal.  
Position  
B to C  
Function  
JMP1  
Normal Operation  
PGA Offset Adjust *  
* Programmable Gain Amplifier requires a special test PROM and is a factory adjustment  
3-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Configuration  
3
3.5 Set Point Error Codes  
Failure of the 943-227-17 preamplifier may result in the following error codes being displayed on the  
controlling device:  
Code E0009 is normally used to identify a Loop communication failure. The error code may be due to a  
baud rate mismatch between the preamplifier and the UDR or 960 system communication board, a failure  
on either the preamplifier or UDR communication board. To clear the error code, the source of the error  
must be corrected.  
3.6 Software Calculations  
The 943-227-17 continuously converts the current output of the detector to a digital value, controls the  
gain of the internal amplifier, performs periodic self checks, communicates with the UDR or Model 960  
controller and provides a radiation value to the UDR of Model 960. An asynchronous, serial  
communication interface, processed through optically isolated drivers/receivers is utilized. The digital  
value is transmitted to the associated UDR or 960 System controller for further processing. Refer to the  
associated controller manual for further information. .  
3-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
(Blank page)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintenance, Calibration and Troubleshooting  
4
Maintenance  
Section 4  
Maintenance, Calibration and Troubleshooting  
4.1 Maintenance  
The 943-227-15 Digital Preamplifier is designed to operate for extended periods of time with no  
scheduled maintenance required. Operation may be verified by periodically actuating the check source  
(provided with the detector and sampling assembly) and observing the response of the unit. If the  
response varies by more than 50% of the normal value, further troubleshooting may be required. If a  
problem develops, troubleshoot the unit per Section 4.3 and the drawings in Appendix B.  
Whenever the preamplifier enclosure is opened, the condition of the internal desiccant package should be  
checked. The desiccant bag is provided in the enclosure to absorb moisture. The desiccant bags should  
be checked for moisture absorption prior to placing into service, or whenever the electronics enclosure  
cover is opened. Each bag contains a color indicator stripe that turns red when it should be replaced.  
4.2 Electronic Calibration  
Electronic Adjustments  
There are no user adjustments available on the 943-227-15 Digital Preamplifier. The Programmable Gain  
Amplifier and Auto Zero are factory adjustments, requiring the use of a factory test PROM and test fixture.  
We strongly recommend the unit be returned to the factory for repair or adjustment.  
Table 4-2 lists the electronic adjustments for the 943-227-15 digital preamplifier. The procedures for  
making these adjustments are contained in the factory Calibration Procedure listed below. Note that  
special test firmware is required to perform the electrical adjustments.  
Table 4-1. Electronic Adjustments  
Potentiometer  
R11  
R12  
Adjustment (943-227-15)  
Programmable Gain Amplifier Offset Adjustment  
Auto Zero Set-Up Adjustment  
Tolerance  
± 2 mV  
± 10 mV  
NOTE  
Refer to the drawings in Appendix B to locate adjustment  
and/or test points, if required.  
The follow factory calibration and test procedures are listed for reference only.  
Document  
Description  
TP943-227-15VL  
TP943-227-15  
TP977-200-15M  
Test Procedure, Victoreen Loop Communications  
Test Procedure, RS232 communications  
Electrometer PCB Test Procedure  
4-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
4.3 Troubleshooting  
WARNING  
Extreme care must be used when troubleshooting a  
system that has power applied.  
troubleshooting precautions apply.  
All standard  
WARNING  
Once a problem has been located, remove all power  
before continuing with the repair.  
CAUTION  
Personnel performing the troubleshooting/repair  
must be qualified to ANSI 45.2.6, 1978, Skill Level II.  
Personnel performing the following procedure must  
be familiar with the operation of the monitoring  
system and the location of each piece of equipment  
used in the system.  
If a problem develops, verify that the voltages at connection point inputs and outputs are present and that  
all wiring is secure. Refer to Appendix B and C for drawings and parts lists. If a PROM requires  
replacement, specify the board revision level when ordering the part.  
Troubleshooting / Adjustments  
Troubleshooting includes a verification of the LED operation status, the DC voltage levels and the loop  
communication loop fuses (Victoreen Loop Communications only).  
Specialized test equipment and firmware are required to adjust the auto-zero and PGA subsystems in the  
electrometer. If the test equipment and firmware is not available, the unit should be returned to Fluke  
Biomedical RMS for re-alignment.  
4-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Connector Designations  
A
Appendix A  
Connector Designations  
A.1 CONNECTOR DESIGNATIONS,  
943-227-15 and 943-227-15VL  
Table A-1. Field Electrical Connections, 943-227-15 and 943-227-15VL  
943-227-15  
943-27-15VL  
Connects To  
977-210-10M  
Connects To  
977-210-10M  
TB1-  
Description  
Description  
+ VL, 15  
VDC  
J4-1  
J3-1  
1
XMIT  
J3-2  
J3-3  
J3-4  
J3-5  
J3-6  
2
3
DTR  
+ TX  
J4-2  
J4-3  
GND  
- TX  
4
REC  
- VL  
J4-4  
Not Used  
Not Used  
+ 15 VDC  
Ground  
- 15 VDC  
Not Used  
Not Used  
5
+ RX  
Not Used  
J3-7  
6
- RX  
7
Not Used  
Ground  
Not Used  
Not Used  
Not Used  
J3-8  
J3-8  
8
9
J3-4  
Not Used  
J3-11  
10  
Not Used  
Detector  
Connections  
Detector  
Connections  
Detector  
From Field  
coaxial cable  
From Field  
coaxial cable  
HV  
Detector MHV  
MHV  
Factory  
Factory  
Terminated,  
To 977-200-  
15M, R41  
Detector  
BNC  
Terminated,  
To 977-200-  
15M, R41  
SIG  
Detector BNC  
A-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
A.2 SWITCH AND JUMPER POSITIONS,  
943-227-15 and 943-227-15VL  
Table A-2. Preamplifier Serial Interface Board 977-210-10M Switch Settings  
Switch Normal Switch Function  
Op.  
SW1  
Open Microprocessor Reset switch, momentary contact  
SW2  
N/A  
Not supplied  
SW3-1  
SW3-2  
SW3-3  
ON  
OFF  
OFF  
OFF Selects 300 baud • 1/ON Selects 4800 baud (normal setting)  
OFF • (No function is implemented in the 977-210-10M)  
OFF Selects Rate Mode (normal) • (Selects Maintenance Mode for  
factory PGA and Auto Zero set-up only)  
SW3-4  
OFF  
OFF • (No function is implemented in the 977-210-10M)  
NOTE  
SW3 is a 4-position DIP switch on the 977-210-10M serial  
interface board.  
Table A-3. Jumper Positions for Preamplifier Serial Interface Board 977-210-10M  
Jumper  
Normal  
Setting  
A to B  
Function  
Alternate Function  
Setting  
JMP1  
B to C  
Enables RS232C  
Enables Victoreen Loop  
(normal)  
JMP2  
N/A  
Deleted, Not Used  
N/A  
Deleted, Not Used  
Table A-4. Jumper Positions for Preamplifier Electrometer Board 977-200-15M  
Jumper  
Normal  
Setting  
A to B  
Function  
Cal.  
Position  
B to C  
Function  
JMP1  
Normal Operation  
PGA Offset Adjust *  
* Programmable Gain Amplifier requires a special test PROM and is a factory adjustment  
A-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Applicable Drawings  
B
Appendix B  
Applicable Drawings  
B.1 Applicable Drawings  
943-227-15 and 943-227-15VL  
Drawing No.  
Description  
943-227-15  
Preamplifier Assembly, RS232  
943-227-15VL  
943-227-15VL-M1  
977-210-10M  
977-210-13M  
977-200-15M  
977-200-18M  
Preamplifier Assembly, Victoreen Loop  
Preamplifier Assembly, Victoreen Loop, w/Conduit Hubs  
Preamplifier HV/Interface Printed Circuit Board Assembly  
Preamplifier Printed Circuit Board Schematic  
Preamplifier Electrometer Board Assembly  
Preamplifier Schematic  
B.2 Related Manuals (Not Supplied with this document)  
SXXXXXX-1  
942A-200C-M1  
943-27-31-1  
Applicable 960 Series system Manual  
Digital Ratemeter  
Current Mode Beta Detector  
B-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
(Blank page)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Bill of Materials  
C
Appendix C  
Bill of Materials  
C.1 Bill of Materials- 943-227-15 and 943-227-15VL  
Part Number  
Description  
943-227-15  
Preamplifier Assembly, RS232  
Preamplifier Assembly, Victoreen Loop  
Preamplifier Assembly, Victoreen Loop, w/Conduit Hubs  
Preamplifier HV/Interface Printed Circuit Board Assembly  
Preamplifier Electrometer Board Assembly  
943-227-15VL  
943-227-15VL-M1  
977-210-10M  
977-200-15M  
C-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
(Blank page)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Cable Termination Instructions  
D
Appendix D  
Cable Termination Instructions  
D.1 Cable Termination Instructions  
The procedures and instructions provided in Appendix D are provided for reference in terminating the field  
cables to the Digital Preamplifier. A mating SHV connector is provided with the preamplifier. Ring lug  
terminals are to be provided by the user. The instructions provided below are based on the use of P/N  
50-100 multi-conductor cable. For other customer-supplied cables, the same basic procedures, modified  
for the specific cable used, may be utilized.  
Detector Cable Termination:  
The detector high voltage and signal cables enter the preamplifier through “Seal Grip” type penetrations  
on the enclosure body. The “Seal Grips” contain a rubber gland that is intended to grip the coaxial cables  
to prevent moisture entry into the enclosure. The “Seal Grips” are an integral part of the detector cables.  
The “Seal Grip” enclosure nut is removed from the fitting. The insulating washer stays with the fitting.  
The detector signal and high voltage cables, with connectors and “Seal Grip” fitting nuts are then inserted  
into the holes preamplifier enclosure. The “Seal Grip: end fittings are then re-installed securely against  
the wall of the enclosure.  
Detector Field Cable Preparation, “MS” Connector:  
Prior to stripping the outer cable jacket, a 4-inch length of 0.75 I.D. shrinkable tubing (provided by the  
user) should be slid over the 0.675 in cable jacket. The shrink tubing will be used to seal and protect the  
end of the cable after the termination process is completed. To prepare the cable for termination, a  
minimum of 7 inches of the cable outer jacket must be stripped off the cable. This will allow for the  
stripping of approximately one inch from the various conductors, and provide 6 inches of actual service  
loop cable. In addition, for EMI/RFI protection, the drain wire must be terminated to an earth ground  
within the enclosure. The length of the drain wire will be based on the distance between the cable entry  
and the ground location. The length of cable jacket to be stripped will then be based on the distance to  
the drain wire grounding point.  
Once the cable outer jacket is stripped to the proper length, the aluminum tape shield and Mylar binder  
may be removed, taking care not to damage the No. 20 AWG drain wire. The drain wire may then be  
separated from the remaining conductors, for routing and termination at the grounding point.  
Next, the HV coaxial cable may be separated from the bundle, followed by the single conductors wires  
that are required for the specific application, and the polyester filler cord. Refer to the project specific  
detector loop drawing, or the generic detector loop drawing provided in Appendix B for actual conductors  
to be used.  
The polyester filler cord may now be removed. Cut the cord as close to the outer jacket as possible.  
The single conductors may now be stripped and soldered to the appropriate connector pin.  
D-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
Detector Field Cable Preparation, 943-227-15VL-M1, Conduit Hub:  
Prior to terminating the cable, the field conduit must be run to the junction box, and the field cable pulled  
into the junction box.  
To prepare the cable for termination, a minimum of 7 inches of the cable outer jacket must be stripped off  
the cable. This will allow for the stripping of approximately one inch from the various conductors, and  
provide 6 inches of actual service loop cable. In addition, for EMI/RFI protection, the drain wire must be  
terminated to an earth ground within the enclosure. The length of the drain wire will be based on the  
distance between the UDR rear panel and the ground location. The length of cable jacket to be stripped  
will then be based on the distance to the drain wire grounding point.  
Once the cable outer jacket is stripped to the proper length, the aluminum tape shield and Mylar binder  
may be removed, taking care not to damage the No. 20 AWG drain wire. The drain wire may then be  
separated from the remaining conductors, for routing and termination at the grounding point.  
Next, the HV coaxial cable may be separated from the bundle, followed by the single conductors wires  
that are required for the specific application, and the polyester filler cord. Refer to the project specific  
detector loop drawing, or the generic detector loop drawing provided in Appendix B for actual conductors  
to be used. The ring lug connectors may now be applied to the single conductor cables, and installed in  
accordance with the project electrical loop diagram.  
Trim each conductor as required to reach its assigned terminal point. Allow a minimum of 1 in. spare  
conductor for future re-termination. Because of the unique requirements of each installation, terminal  
lugs for the detector single conductor terminations are not included, and are supplied by the user. Ring  
terminal, Crimp type lugs, Burndy YAE type, sized for the 18 AWG wire, or equivalent, are recommended.  
The terminal block is designed to accept a maximum 14 AWG conductor, and is provided with 6-32  
mounting screws. Strip the conductor as required by the lug manufacturer, install the lug, and terminate  
the conductor to the appropriate terminal. The screw terminals may now be torqued to 12 in-lbs.  
The mating HV connector, P/N 30-92-1 (Kings 1705-14), may now be terminated per the applicable Kings  
data sheet attached.  
The polyester filler cord may now be removed. Cut the cord as close to the outer jacket as possible  
D-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Cable Termination Instructions  
D
HV CONNECTOR NOTES (HV)  
The mating HV connector, P/N 30-92-1 (Kings 1705-14), may now be terminated per the applicable Kings  
data sheet attached  
LOCATION  
P/N  
DOCUMENT  
DESCRIPTION  
SHV  
SHV  
92-9105-A  
(1705-1)  
CP-400, Kings  
Cabling  
Procedure, Trim Code  
441, 442  
High Voltage Connector,  
SHV, male crimp contact  
pin (Use Tool KTH-100 with  
KTH-2062)  
SHV  
Alternate  
AMP Instruction  
Sheet  
408-2187  
High Voltage Connector,  
SHV, male crimp contact  
pin (Use Tool 69710-1 with  
220028-2 Die or Integral  
Die Tool 220022-2)  
Connector  
92-9105-A  
(51426-5)  
SHV  
32-92-1  
(1705-14)  
CP-1001, Kings Cabling  
Procedure, CP-1000,  
High Voltage Connector, SHV,  
Female solder contact pin  
D-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
D-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Cable Termination Instructions  
D
D-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
D-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Cable Termination Instructions  
D
D-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
943-227-15 Digital Preamplifier  
Operator Manual  
D-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Modifications  
E
Appendix E  
Modification Sheets, Engineering Instructions  
E.1 Modification Sheets, Engineering Instructions  
P/N  
Description  
943-227-15VL-M1  
Modification consists of replacing the standard 8in  
H x 10in W x 4in D Enclosure with a 8in H x 10in W  
x 6in D Enclosure and replacing the field cable  
“Seal Grip” type entry fitting with a 2.0 in diameter  
sealed conduit fitting.  
E-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
(Blank page)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
(Blank page)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke Biomedical  
6045 Cochran Road  
Cleveland, Ohio 44139  
440.498.2564  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Euphonix Range 740006 User Manual
Eureka Vacuum Cleaner 79A User Manual
Extron electronic Stereo Amplifier PA 300 User Manual
Fostex DVD Recorder PD204 User Manual
Freud Tools Router FT3000VCE User Manual
Fujitsu Server M10 4 User Manual
Gaggenau Dishwasher DF 241 User Manual
GE Dishwasher GDWF100R User Manual
GE Hot Beverage Maker 169040 User Manual
GE Mixer 169160 User Manual