Fluke Power Supply 80 Series III User Manual

®
80 Series III  
Multimeters  
Users Manual  
October 1997 Rev.5, 12/03  
1997-2003 Fluke Corporation, All rights reserved.  
All product names are trademarks of their respective companies.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Title  
Introduction....................................................................................................................  
Safety Information .........................................................................................................  
Your Meter’s Features...................................................................................................  
1
1
4
Power-Up Options .................................................................................................... 11  
Automatic Power-Off................................................................................................. 11  
Input Alert™ Feature ................................................................................................ 12  
Making Measurements .................................................................................................. 12  
Measuring AC and DC Voltage................................................................................. 12  
Testing for Continuity................................................................................................ 14  
Measuring Resistance .............................................................................................. 16  
Using Conductance for High Resistance or Leakage Tests ..................................... 18  
Measuring Capacitance............................................................................................ 18  
Testing Diodes.......................................................................................................... 21  
Measuring AC or DC Current.................................................................................... 22  
Measuring Frequency............................................................................................... 25  
Measuring Duty Cycle............................................................................................... 27  
Determining Pulse Width.......................................................................................... 28  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
80 Series III  
Users Manual  
Analog Bar Graph.......................................................................................................... 28  
Model 87 Bar Graph.................................................................................................. 28  
Models 83 and 85 Bar Graph.................................................................................... 29  
4-1/2 Digit Mode (Model 87) .......................................................................................... 29  
MIN MAX Recording Mode ............................................................................................ 30  
®
Touch Hold Mode....................................................................................................... 32  
Relative Mode................................................................................................................ 32  
Zoom Mode (Models 83 and 85)............................................................................... 32  
Uses for the Zoom Mode (Models 83 and 85)........................................................... 33  
Maintenance .................................................................................................................. 33  
General Maintenance................................................................................................ 33  
Testing the Fuses...................................................................................................... 34  
Replacing the Battery................................................................................................ 35  
Replacing the Fuses ................................................................................................. 35  
Service and Parts........................................................................................................... 36  
Specifications................................................................................................................. 41  
ii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table  
Title  
1.  
2.  
3.  
4.  
International Electrical Symbols.........................................................................................  
Inputs .................................................................................................................................  
Rotary Switch Positions .....................................................................................................  
Pushbuttons .......................................................................................................................  
Display Features ................................................................................................................  
2
4
5
6
9
5.  
6.  
7.  
8.  
9.  
Estimating Capacitance Values Over 5 Microfarads.......................................................... 20  
Functions and Trigger Levels for Frequency Measurements............................................. 26  
MIN MAX Functions ........................................................................................................... 31  
Replacement Parts............................................................................................................. 38  
Accessories........................................................................................................................ 40  
Models 85 and 87 AC Voltage Function Specifications...................................................... 42  
Model 83 AC Voltage Function Specifications ................................................................... 43  
DC Voltage, Resistance, and Conductance Function Specifications................................. 44  
Current Function Specifications ......................................................................................... 45  
Capacitance and Diode Function Specifications................................................................ 47  
Frequency Counter Specifications ..................................................................................... 47  
Frequency Counter Sensitivity and Trigger Levels............................................................. 48  
Electrical Characteristics of the Terminals ......................................................................... 49  
MIN MAX Recording Specifications ................................................................................... 50  
10.  
11.  
12.  
13.  
14.  
15.  
16.  
17.  
18.  
19.  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
80 Series III  
Users Manual  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure  
Title  
Page  
1. Display Features (Model 87 Shown).......................................................... 8  
2. Measuring AC and DC Voltage.................................................................. 13  
3. Testing for Continuity................................................................................. 15  
4. Measuring Resistance............................................................................... 17  
5. Measuring Capacitance............................................................................. 19  
6. Testing a Diode ......................................................................................... 21  
7. Measuring Current..................................................................................... 23  
8. Components of Duty Cycle Measurements ............................................... 27  
9. Testing the Current Fuses......................................................................... 34  
10. Battery and Fuse Replacement ................................................................. 37  
11. Replaceable Parts..................................................................................... 39  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
80 Series III  
Users Manual  
vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
In this manual, a Warning identifies conditions and  
actions that pose hazards to the user. A Caution  
identifies conditions and actions that may damage the  
meter or the equipment under test.  
Introduction  
WWarning  
Read "Safety Information" before you use the  
meter.  
International symbols used on the meter and in this  
manual are explained in Table 1.  
Except where noted, the descriptions and instructions in  
this manual apply to Series III Models 83, 85, 87, and  
87/E multimeters. Model 87 is shown in all illustrations.  
WWarning  
To avoid possible electric shock or personal  
injury, follow these guidelines:  
Safety Information  
Do not use the meter if it is damaged.  
Before you use the meter, inspect the  
case. Look for cracks or missing plastic.  
Pay particular attention to the insulation  
surrounding the connectors.  
This meter complies with:  
EN61010.1:1993  
ANSI/ISA S82.01-1994  
CAN/CSA C22.2 No. 1010.1-92  
1000 V Overvoltage Category III, Pollution Degree 2  
UL3111-1  
Make sure the battery door is closed and  
latched before you operate the meter.  
Use the meter only as specified in this manual, otherwise  
the protection provided by the meter may be impaired.  
Replace the battery as soon as the  
battery indicator (M) appears.  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Table 1. International Electrical Symbols  
AC (Alternating Current)  
Earth ground  
DC (Direct Current)  
AC or DC  
Fuse  
Conforms to European Union directives  
Refer to the manual for information  
about this feature.  
Conforms to relevant Canadian  
Standards Association directives  
Battery  
Double insulated  
Inspected and licensed by TÜV Product Services.  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Safety Information  
Caution  
Remove test leads from the meter before  
you open the battery door.  
To avoid possible damage to the meter or to  
the equipment under test, follow these  
guidelines:  
Inspect the test leads for damaged  
insulation or exposed metal. Check the  
test leads for continuity. Replace  
damaged test leads before you use the  
meter.  
Disconnect circuit power and discharge  
all high-voltage capacitors before testing  
resistance, continuity, diodes, or  
capacitance.  
Do not use the meter if it operates  
abnormally. Protection may be impaired.  
When in doubt, have the meter serviced.  
Use the proper terminals, function, and  
range for your measurements.  
Do not operate the meter around  
explosive gas, vapor, or dust.  
Before measuring current, check the  
meter’s fuses. (See "Testing the Fuses".)  
Use only a single 9 V battery, properly  
installed in the meter case, to power the  
meter.  
When servicing the meter, use only  
specified replacement parts.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
80 Series III  
Users Manual  
To protect yourself, use the following guidelines:  
Table 2. Inputs  
Description  
Use caution when working with voltages above 30 V  
ac rms, 42 V ac peak, or 60 V dc. Such voltages  
pose a shock hazard.  
Terminal  
Page  
A
Input for 0 A to 10.00 A  
current measurements  
22  
22  
NA  
When using the probes, keep your fingers behind the  
finger guards.  
mA µA Input for 0 µA to 400 mA  
current measurements  
Connect the common test lead before you connect  
the live test lead. When you disconnect test leads,  
disconnect the live test lead first.  
COM  
Return terminal for all  
measurements  
Avoid working alone.  
V eG Input for voltage,  
continuity, resistance,  
diode, capacitance,  
V: 12  
e: 16  
G: 21  
E:18  
When measuring current, turn off circuit power  
before connecting the meter in the circuit. Remember  
to place the meter in series with the circuit.  
frequency, and duty  
cycle measurements  
Frequency: 25  
Duty cycle: 27  
Your Meter’s Features  
Tables 2 through 5 briefly describe your meter’s features  
and give page numbers where you can find more detailed  
information about the features.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Your Meter’s Features  
Table 3. Rotary Switch Positions  
Function  
Switch Position  
Page  
12  
K
L
AC voltage measurement  
DC voltage measurement  
400 mV dc voltage range  
12  
12  
d
mV  
ReE  
R Continuity test  
14  
16  
18  
21  
22  
e Resistance measurement  
E Capacitance measurement  
Diode test  
G
mA  
A
DC or AC current measurements from 0 mA to 10.00 A  
µA  
DC or AC current measurements from 0 µA to 4000 µA  
22  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Table 4. Pushbuttons  
Button Function  
Button  
Function  
ReE  
Selects capacitance.  
18  
22  
30  
U
mA/A, µA  
Power-up  
Switches between dc and ac current.  
Disables automatic power-off feature.  
(Blue  
button)  
Any switch  
position  
Starts recording of minimum and maximum values. Steps the display through  
MIN, MAX, AVG (average), and present readings.  
M
K
Power-up  
Enables high-accuracy 1-second response time for MIN MAX recording.  
30  
Any switch  
position  
Switches between the ranges available for the selected function. To return to  
autoranging, hold the button down for 1 second.  
See ranges in  
specifications.  
®
Manually selecting a range causes the meter to exit the Touch Hold , MIN  
MAX, and REL (relative) modes.  
Power-up  
For servicing purposes only.  
NA  
32  
Any switch  
position  
Touch Hold captures the present reading on the display. When a new, stable  
reading is detected, the meter beeps and displays the new reading.  
I
MIN MAX  
recording  
Stops and starts recording without erasing recorded values.  
30  
25  
Frequency  
counter  
Stops and starts the frequency counter.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Your Meter’s Features  
Table 4. Pushbuttons (cont)  
Button  
Function  
Button Function  
Turns the backlight on and off.  
Page  
Any switch  
position  
b Model 87:  
yellow button  
For Model 87, hold the yellow button down for one second to enter the  
4-1/2 digit mode. To return to the 3-1/2 digit mode, hold the button down  
only until all display segments turn on (about one second).  
29  
b Models 83,  
85: gray  
button  
Continuity  
ReE  
Turns the continuity beeper on and off.  
14  
30  
NA  
32  
T
MIN MAX  
recording  
On Model 87, switches between 250 µs and 100 ms or 1 s response  
times.  
Power-up  
Disables the beeper for all functions.  
C
(Relative  
mode)  
Any switch  
position  
Stores the present reading as a reference for subsequent readings. The  
display is zeroed, and the stored reading is subtracted from all  
subsequent readings.  
Power-up  
For Models 83 and 85, enables zoom mode for the bar graph.  
Starts the frequency counter.  
32  
25  
27  
NA  
Any switch  
position  
F
Press again to enter duty cycle mode.  
Power-up  
Provides >4000 Minput impedance for the 400 mV dc range.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
80 Series III  
Users Manual  
9
8
6
7
5
4
110  
3
2
11  
1
12  
13  
iy1f.eps  
Figure 1. Display Features (Model 87 Shown)  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Your Meter’s Features  
Table 5. Display Features  
Indication  
Number  
Feature  
±
A
Polarity indicator for the analog bar graph.  
Relative (REL) mode is active.  
28  
B
32  
Q
C
D
The continuity beeper is on.  
S
Indicates negative readings. In relative mode, this sign indicates that the  
present input is less than the stored reference.  
32  
-
E
F
The battery is low. WWarning: To avoid false readings, which could lead  
to possible electric shock or personal injury, replace the battery as soon  
as the battery indicator appears.  
35  
The meter is in autorange mode and automatically selects the range with the  
best resolution.  
NA  
AUTO  
100 ms  
MAX MIN AVG  
G
H
I
Indicators for minimum-maximum recording mode.  
30  
32  
Touch Hold is active.  
AC DC  
Indicator for ac or dc voltage or current. AC voltage and current is displayed as  
an rms (root mean square) value.  
12, 22  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Table 5. Display Features (continued)  
Indication  
Number  
Feature  
J
A, µA, mA  
A: Amperes (amps). The unit of current.  
µA: Microamp. 1 x 10-6 or 0.000001 amperes.  
mA: Milliamp. 1 x 10-3 or 0.001 amperes.  
22  
V, mV  
V: Volts. The unit of voltage.  
12  
18  
mV: Millivolt. 1 x 10-3 or 0.001 volts.  
µF, nF  
F: Farad. The unit of capacitance.  
µF: Microfarad. 1 x 10-6 or 0.000001 farads.  
nF: Nanofarad. 1 x 10-9 or 0.000000001 farads.  
nS  
S: Siemen. The unit of conductance.  
nS: Nanosiemen. 1 x 10-9 or 0.000000001 siemens.  
18  
%
Percent. Used for duty cycle measurements.  
27  
16  
e, Me, ke  
: Ohm. The unit of resistance.  
M: Megohm. 1 x 106 or 1,000,000 ohms.  
k: Kilohm. 1 x 103 or 1000 ohms.  
Hz, kHz, MHz Hz: Hertz. The unit of frequency.  
kHz: Kilohertz. 1 x 103 or 1000 hertz.  
25  
MHz: Megahertz. 1 x 106 or 1,000,000 hertz.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Your Meter’s Features  
Table 5. Display Features (continued)  
Number  
Feature  
Indication  
Page  
K
4000 mV  
Displays the currently selected range.  
See specifications  
for ranges for each  
function.  
L
M
Analog bar graph  
Provides an analog indication of the present inputs.  
28  
The input (or the relative value when in relative mode) is too large  
for the selected range. For duty cycle measurements OL is  
displayed when the input signal stays high or low.  
Duty cycle: 27  
0L  
Power-Up Options  
Automatic Power-Off  
Holding a button down while turning the meter on  
activates a power-up option. Table 4 includes the power-  
up options available. These options are also listed on the  
back of the meter.  
The meter automatically turns off if you do not turn the  
rotary switch or press a button for 30 minutes. To disable  
automatic power-off, hold down the blue button while  
turning the meter on. Automatic power-off is always  
disabled in MIN MAX recording mode.  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Input Alert™ Feature  
Measuring AC and DC Voltage  
If a test lead is plugged into the mA/µA or A terminal, but  
the rotary switch is not correctly set to the mA/µA or A  
position, the beeper warns you by making a chirping  
sound. This warning is intended to stop you from  
attempting to measure voltage, continuity, resistance,  
capacitance, or diode values when the leads are plugged  
into a current terminal. Placing the probes across (in  
parallel with) a powered circuit when a lead is plugged into  
a current terminal can damage the circuit you are testing  
and blow the meter’s fuse. This can happen because the  
resistance through the meter’s current terminals is very  
low, so the meter acts like a short circuit.  
Voltage is the difference in electrical potential between  
two points. The polarity of ac (alternating current) voltage  
varies over time, while the polarity of dc (direct current)  
voltage is constant over time. The meter presents ac  
voltage values as rms (root mean square) readings. The  
rms value is the equivalent dc voltage that would produce  
the same amount of heat in a resistance as the measured  
sinewave voltage. Models 85 and 87 feature true rms  
readings, which are accurate for other wave forms (with  
no dc offset) such as square waves, triangle waves, and  
staircase waves.  
The meter’s voltage ranges are 400 mV, 4 V, 40 V, 400 V,  
and 1000 V. To select the 400 mV dc range, turn the  
rotary switch to mV.  
Making Measurements  
The following sections describe how to take  
measurements with your meter.  
To measure ac or dc voltage, set up and connect the  
meter as shown in Figure 2.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Making Measurements  
The following are some tips for measuring voltage:  
AC Voltage  
When you measure voltage, the meter acts  
approximately like a 10 M(10,000,000 )  
impedance in parallel with the circuit. This loading  
effect can cause measurement errors in high-  
impedance circuits. In most cases, the error is  
negligible (0.1% or less) if the circuit impedance is  
10 k(10,000 ) or less.  
RMS MULTIMETER  
87III TRUE  
MIN MAX  
RANGE  
HOLD  
Hz  
H
Switch Box  
REL  
41/2 DIGITS  
PEAK MIN MAX  
1
Second  
mV  
mA  
A
V
V
µA  
V
OFF  
A
mA µA COM  
V
For better accuracy when measuring the dc offset of  
an ac voltage, measure the ac voltage first. Note the  
ac voltage range, then manually select a dc voltage  
range equal to or higher than the ac range. This  
procedure improves the accuracy of the dc  
measurement by ensuring that the input protection  
circuits are not activated.  
400mA MAX  
FUSED  
10A MAX  
FUSED  
X  
!
DC Voltage  
RMS MULTIMETER  
87III TRUE  
MIN MAX  
RANGE  
HOLD  
Hz  
H
REL  
PEAK MIN MAX  
41/2 DIGITS  
1
Second  
mV  
mA  
A
V
V
µA  
V
OFF  
A
mA µA COM  
V
+
400mA MAX  
FUSED  
10A MAX  
FUSED  
X  
!
iy2f.eps  
Figure 2. Measuring AC and DC Voltage  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
The continuity function detects intermittent opens and  
shorts lasting as little as 1 millisecond (0.001 second).  
These brief contacts cause the meter to emit a short beep.  
Testing for Continuity  
Caution  
To avoid possible damage to the meter or to  
the equipment under test, disconnect circuit  
power and discharge all high-voltage  
capacitors before testing for continuity.  
Continuity is the presence of a complete path for current  
flow. The continuity test features a beeper that sounds if a  
circuit is complete. The beeper allows you to perform  
quick continuity tests without having to watch the display.  
To test for continuity, set up the meter as shown in  
Figure 3.  
Press Tto turn the continuity beeper on or off.  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Making Measurements  
For in-circuit tests, turn circuit power off.  
Activates  
RMS MULTIMETER  
87III TRUE  
RMS MULTIMETER  
87III TRUE  
continuity  
beeper  
MIN MAX  
RANGE  
HOLD  
Hz  
H
MIN MAX  
RANGE  
REL  
D  
Hz  
H
REL  
4
1/2 DIGITS  
PEAK MIN MAX  
4
1
1/2 DIGITS PEAK MIN MAX  
conds  
Seconds  
mV  
mV  
mA  
A
mA  
A
V
V
V
V
µA  
µA  
OFF  
OFF  
A
mA µA COM  
V
400mA MAX  
FUSED  
400mA MAX  
FUSED  
10A MAX  
FUSED  
10A MAX  
FUSED  
1
!
!
OFF  
(open)  
ON  
(closed)  
iy4f.eps  
Figure 3. Testing for Continuity  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
The following are some tips for measuring resistance:  
Measuring Resistance  
Because the meter’s test current flows through all  
possible paths between the probe tips, the measured  
value of a resistor in a circuit is often different from  
the resistor’s rated value.  
Caution  
To avoid possible damage to the meter or to  
the equipment under test, disconnect circuit  
power and discharge all high-voltage  
capacitors before measuring resistance.  
The test leads can add 0.1 to 0.2 of error to  
resistance measurements. To test the leads, touch  
the probe tips together and read the resistance of the  
leads. If necessary, you can use the relative (REL)  
mode to automatically subtract this value.  
Resistance is an opposition to current flow. The unit of  
resistance is the ohm (). The meter measures resistance  
by sending a small current through the circuit. Because  
this current flows through all possible paths between the  
probes, the resistance reading represents the total  
resistance of all paths between the probes.  
The resistance function can produce enough voltage  
to forward-bias silicon diode or transistor junctions,  
causing them to conduct. To avoid this, do not use  
the 40 Mrange for in-circuit resistance  
measurements.  
The meter’s resistance ranges are 400 , 4 k, 40 k,  
400 k, 4 M, and 40 M.  
To measure resistance, set up the meter as shown in  
Figure 4.  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Making Measurements  
In-Circuit Resistance Measurements  
Isolating a Potentiometer  
Circuit Power  
1
OFF  
2
3
Disconnect  
2
1
RMS MULTIMETER  
87III TRUE  
3
H
MIN MAX  
RANGE  
HOLD  
Hz  
Isolating a Resistor  
REL  
PEAK MIN MAX  
4
1
1/2 DIGITS  
Seconds  
mV  
mA  
A
V
V
µA  
OFF  
A
mA µA  
COM  
V
400mA MAX  
FUSED  
10A MAX  
FUSED  
X  
!
Disconnect  
iy6f.eps  
Figure 4. Measuring Resistance  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
The following are some tips for measuring conductance:  
Using Conductance for High Resistance or  
Leakage Tests  
High-resistance readings are susceptible to electrical  
noise. To smooth out most noisy readings, enter the  
MIN MAX recording mode; then scroll to the average  
(AVG) reading.  
Conductance, the inverse of resistance, is the ability of a  
circuit to pass current. High values of conductance  
correspond to low values of resistance.  
The unit of conductance is the Siemen (S). The meter’s  
40 nS range measures conductance in nanosiemens  
(1 nS = 0.000000001 Siemens). Because such small  
amounts of conductance correspond to extremely high  
resistance, the nS range lets you determine the  
resistance of components up to 100,000 M, or  
100,000,000,000 (1/1 nS = 1,000 M).  
There is normally a residual conductance reading  
with the test leads open. To ensure accurate  
readings, use the relative (REL) mode to subtract the  
residual value.  
Measuring Capacitance  
Caution  
To measure conductance, set up the meter as shown for  
measuring resistance (Figure 4); then press Kuntil  
the nS indicator appears on the display.  
To avoid possible damage to the meter or to  
the equipment under test, disconnect circuit  
power and discharge all high-voltage  
capacitors before measuring capacitance.  
Use the dc voltage function to confirm that  
the capacitor is discharged.  
Capacitance is the ability of a component to store an  
electrical charge. The unit of capacitance is the farad (F).  
Most capacitors are in the nanofarad to microfarad range.  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Making Measurements  
The meter measures capacitance by charging the  
capacitor with a known current for a known period of time,  
measuring the resulting voltage, then calculating the  
capacitance. The measurement takes about 1 second per  
range. The capacitor charge can be up to 1.2 V.  
RMS MULTIMETER  
87IIITRUE  
Select  
Capacitance  
µ
nF  
MIN MAX  
RANGE  
HOLD  
Hz  
H
REL  
4
1/2 DIGITS PEAK MIN MAX  
1
Seconds  
mV  
mA  
A
V
V
The meter’s capacitance ranges are 5 nF, 0.05 µF,  
0.5 µF, and 5 µF.  
µA  
OFF  
A
mA µA  
COM  
V
To measure capacitance, set up the meter as shown in  
Figure 5.  
400
F
CAT II  
10A MAX  
FUSED  
1000V MAX  
The following are some tips for measuring capacitance:  
To speed up measurements of similar values, press  
Kto manually select the proper range.  
To improve the accuracy of measurements less than  
5 nF, use the relative (REL) mode to subtract the  
residual capacitance of the meter and leads.  
+
+
+
+
+
+
+
+
+
iy10f.eps  
Figure 5. Measuring Capacitance  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
To estimate capacitance values above 5 µF, use the  
current supplied by the meter’s resistance function,  
as follows:  
Table 6. Estimating Capacitance Values Over  
5 Microfarads  
µF/second  
of Charge  
1. Set up the meter to measure resistance.  
Suggested  
Range*  
Expected Capacitance  
Up to 10 µF  
Time  
2. Press Kto select a range based on the  
value of capacitance you expect to measure  
(refer to Table 6.)  
4 Me  
400 ke  
40 ke  
4 ke  
0.3  
3
11 µF to 100 µF  
3. Discharge the capacitor.  
101 µF to 1000 µF  
1001 µF to 10,000 µF  
10,000 µF to 100,000 µF  
30  
4. Place the meter’s leads across the capacitor;  
then time how long it takes for the display to  
reach OL.  
300  
3000  
400 e  
5. Multiply the charge time from step 4 by the  
appropriate value in the µF/second of Charge  
Time column in 6. The result is the estimated  
capacitance value in microfarads (µF).  
*These ranges keep the full-charge time between 3.7 seconds  
and 33.3 seconds for the expected capacitance values. If the  
capacitor charges too quickly for you to time, select the next  
higher resistance range.  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Making Measurements  
Testing Diodes  
Forward Bias  
Caution  
RMS MULTIMETER  
87III TRUE  
Typical  
Reading  
To avoid possible damage to the meter or to  
the equipment under test, disconnect circuit  
power and discharge all high-voltage  
capacitors before testing diodes.  
+
MIN MAX  
RANGE  
HOLD  
Hz  
H
REL  
PEAK MIN MAX  
4
1/2 DIGITS  
1
Seconds  
mV  
mA  
A
V
V
µA  
OFF  
Use the diode test to check diodes, transistors, silicon  
controlled rectifiers (SCRs), and other semiconductor  
devices. This function tests a semiconductor junction by  
sending a current through the junction, then measuring  
the junction’s voltage drop. A good silicon junction drops  
between 0.5 V and 0.8 V.  
A
mA µA  
COM  
V
400mA MAX  
FUSED  
10A MAX  
1
FUSED  
!
Reverse Bias  
To test a diode out of a circuit, set up the meter as shown  
in Figure 6. For forward-bias readings on any  
semiconductor component, place the red test lead on the  
component’s positive terminal and place the black lead  
on the component’s negative terminal.  
RMS MULTIMETER  
87III TRUE  
+
MIN MAX  
RANGE  
HOLD  
Hz  
H
REL  
4
1/2 DIGITS PEAK MIN MAX  
1
Seconds  
mV  
mA  
A
V
V
µA  
OFF  
In a circuit, a good diode should still produce a forward-  
bias reading of 0.5 V to 0.8 V; however, the reverse-bias  
reading can vary depending on the resistance of other  
pathways between the probe tips.  
A
mA µA  
COM  
V
400mA MAX  
FUSED  
10A MAX  
FUSED  
!
iy9f.eps  
Figure 6. Testing a Diode  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
The meter’s current ranges are 400 µA, 4000 µA,  
40 mA, 400 mA, 4000 mA, and 10 A. AC current is  
displayed as an rms value.  
Measuring AC or DC Current  
WWarning  
Never attempt an in-circuit current  
To measure current, refer to Figure 7 and proceed as  
follows:  
measurement where the open-circuit  
potential to earth is greater than 1000 V. You  
may damage the meter or be injured if the  
fuse blows during such a measurement.  
1. Turn off power to the circuit. Discharge all high-  
voltage capacitors.  
2. Insert the black lead into the COM terminal. For  
currents between 4 mA and 400 mA, insert the red  
lead into the mA/µA terminal. For currents above  
400 mA, insert the red lead into the A terminal.  
Caution  
To avoid possible damage to the meter or to  
the equipment under test, check the meter’s  
fuses before measuring current. Use the  
proper terminals, function, and range for  
your measurement. Never place the probes  
across (in parallel with) any circuit or  
component when the leads are plugged into  
the current terminals.  
Note  
To avoid blowing the meter’s 400 mA fuse, use  
the mA/µA terminal only if you are sure the  
current is less than 400 mA.  
Current is the flow of electrons through a conductor. To  
measure current, you must break the circuit under test,  
then place the meter in series with the circuit.  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Making Measurements  
Total current to circuit  
Circuit Power:  
OFF to connect meter.  
ON for measurement.  
OFF to disconnect meter.  
1
4
RMS MULTIMETER  
87III TRUE  
5
AC DC  
H
MIN MAX  
RANGE  
HOLD  
Hz  
3
mA  
A
Current through one component  
REL  
4
1/2 DIGITS PEAK MIN MAX  
1
Seconds  
mV  
µA  
mA  
A
V
V
µA  
2
OFF  
mA µA  
COM  
V
A
MAX  
D  
X  
10A MAX  
FUSED  
5
iy7f.eps  
Figure 7. Measuring Current  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
3. If you are using the A terminal, set the rotary switch to  
mA/A. If you are using the mA/µA terminal, set the  
rotary switch to µA for currents below 4000 µA  
(4 mA), or mA/A for currents above 4000 µA.  
The following are some tips for measuring current:  
If the current reading is 0 and you are sure the meter  
is set up correctly, test the meter’s fuses as described  
under "Testing the Fuses".  
4. To measure ac current, press the blue button.  
A current meter drops a small voltage across itself,  
which might affect circuit operation. You can calculate  
this burden voltage using the values listed in the  
specifications in Table 14.  
5. Break the circuit path to be tested. Touch the black  
probe to the more negative side of the break; touch  
the red probe to the more positive side of the break.  
Reversing the leads will produce a negative reading,  
but will not damage the meter.  
6. Turn on power to the circuit; then read the display. Be  
sure to note the unit given at the right side of the  
display (µA, mA, or A).  
7. Turn off power to the circuit and discharge all high-  
voltage capacitors. Remove the meter and restore the  
circuit to normal operation.  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Making Measurements  
The following are some tips for measuring frequency:  
Measuring Frequency  
Frequency is the number of cycles a signal completes  
each second. The meter measures the frequency of a  
voltage or current signal by counting the number of times  
the signal crosses a threshold level each second.  
If a reading shows as 0 Hz or is unstable, the input  
signal may be below or near the trigger level. You can  
usually correct these problems by selecting a lower  
range, which increases the sensitivity of the meter. In  
the L function, the lower ranges also have lower  
trigger levels.  
Table 7 summarizes the trigger levels and applications for  
measuring frequency using the various ranges of the  
meter’s voltage and current functions.  
If a reading seems to be a multiple of what you  
expect, the input signal may be distorted. Distortion  
can cause multiple triggerings of the frequency  
counter. Selecting a higher voltage range might solve  
this problem by decreasing the sensitivity of the  
meter. You can also try selecting a dc range, which  
raises the trigger level. In general, the lowest  
frequency displayed is the correct one.  
To measure frequency, connect the meter to the signal  
source; then press F. Pressing Tswitches the  
trigger slope between + and -, as indicated by the symbol  
at the left side of the display (refer to Figure 8 under  
"Measuring Duty Cycle"). Pressing Istops and  
starts the counter.  
The meter autoranges to one of five frequency ranges:  
199.99 Hz, 1999.9 Hz, 19.999 kHz, 199.99 kHz, and  
greater than 200 kHz. For frequencies below 10 Hz, the  
display is updated at the frequency of the input. Between  
0.5 Hz and 0.3 Hz, the display may be unstable. Below  
0.3 Hz, the display shows 0.000 Hz.  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Table 7. Functions and Trigger Levels for Frequency Measurements  
Approximate  
Function  
Range  
Trigger Level  
Typical Application  
K
4 V, 40 V, 400 V, 0 V  
1000 V  
Most signals.  
K
400 mV  
0 V  
High-frequency 5 V logic signals. (The dc-coupling of the L function can  
attenuate high-frequency logic signals, reducing their amplitude enough  
to interfere with triggering.)  
L
L
L
L
L
400 mV  
4 V  
40 mV  
1.7 V  
4 V  
Refer to the measurement tips given before this table.  
5 V logic signals (TTL).  
40 V  
Automotive switching signals.  
400 V  
1000 V  
40 V  
Refer to the measurement tips given before this table.  
400 V  
ReEG Frequency counter characteristics are not specified for these functions.  
\
µAF  
^
AF  
All ranges  
0 A  
AC current signals.  
400 µA  
40 mA  
4 A  
Refer to the measurement tips given before this table.  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Making Measurements  
frequency function, you can change the slope for the  
meter’s counter by pressing T.  
Measuring Duty Cycle  
Duty cycle (or duty factor) is the percentage of time a  
signal is above or below a trigger level during one cycle  
(Figure 8). The duty cycle mode is optimized for  
measuring the on or off time of logic and switching  
signals. Systems such as electronic fuel injection systems  
and switching power supplies are controlled by pulses of  
varying width, which can be checked by measuring duty  
cycle.  
For 5 V logic signals, use the 4 V dc range. For 12 V  
switching signals in automobiles, use the 40 V dc range.  
For sine waves, use the lowest range that does not result  
in multiple triggering. (Normally, a distortion-free signal  
can be up to ten times the amplitude of the selected  
voltage range.)  
If a duty cycle reading is unstable, press MIN MAX; then  
scroll to the AVG (average) display.  
To measure duty cycle, set up the meter to measure  
frequency; then press Hz a second time. As with the  
+Slope  
-Slope  
Trigger Point  
Trigger Point  
30% Above  
+Slope  
70% Below  
-Slope  
100%  
iy3f.eps  
Figure 8. Components of Duty Cycle Measurements  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Determining Pulse Width  
Analog Bar Graph  
For a periodic waveform (its pattern repeats at equal time  
intervals), you can determine the amount of time that the  
signal is high or low as follows:  
The analog bar graph functions like the needle on an  
analog meter, but without the overshoot. The bar graph is  
updated 40 times per second. Because the graph  
responds 10 times faster than the digital display, it is  
useful for making peak and null adjustments and  
observing rapidly changing inputs.  
1. Measure the signal’s frequency.  
2. Press Fa second time to measure the signal’s  
duty cycle. Press T to select a measurement of  
the signal’s positive or negative pulse. (Refer to  
Figure 8.)  
Model 87 Bar Graph  
Model 87’s bar graph consists of 32 segments. The  
position of the pointer on the display represents the last  
three digits of the digital display. For example, for inputs  
of 500 , 1500 , and 2500 , the pointer is near 0.5 on  
the scale. If the last three digits are 999, the pointer is at  
the far right of the scale. As the digits increment past 000,  
the pointer wraps back to the left side of the display. The  
polarity indicator at the left of the graph indicates the  
polarity of the input.  
3. Use the following formula to determine the pulse  
width:  
Pulse Width = % Duty Cycle ÷ 100  
(in seconds)  
Frequency  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
4-1/2 Digit Mode (Model 87)  
Models 83 and 85 Bar Graph  
4-1/2 Digit Mode (Model 87)  
The bar graph on Models 83 and 85 consists of 43  
segments. The number of lit segments is relative to the  
full-scale value of the selected range. The polarity  
indicator at the left of the graph indicates the polarity of  
the input. For example, if the 40 V range is selected, the  
"4" on the scale represents 40 V. An input of -30 V would  
light the negative sign and the segments up to the "3" on  
the scale.  
On a Model 87 meter, pressing the yellow button for one  
second causes the meter to enter the high-resolution,  
4-1/2 digit mode. Readings are displayed at 10 times the  
normal resolution with a maximum display of 19,999  
counts. The display is updated once per second. The  
4-1/2 digit mode works in all modes except capacitance  
and the 250 µs and 100 ms MIN MAX modes.  
To return to the 3-1/2 digit mode, press the yellow button  
only until all of the display segments turn on (about one  
second).  
If the input equals or exceeds the 4096 counts on a  
manually-selected range, all segments are lit and®  
appears to the right of the bar graph. The graph does not  
operate with the capacitance or frequency counter  
functions.  
The bar graph on Models 83 and 85 also has a zoom  
function, as described under "Zoom Mode".  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
The 100 millisecond response time is best for recording  
power supply surges, inrush currents, and finding  
intermittent failures. This response time follows the  
update time of the analog display.  
MIN MAX Recording Mode  
The MIN MAX mode records minimum and maximum  
input values. When the inputs go below the recorded  
minimum value or above the recorded maximum value,  
the meter beeps and records the new value. This mode  
can be used to capture intermittent readings, record  
maximum readings while you are away, or record  
readings while you are operating the equipment under  
test and cannot watch the meter. MIN MAX mode can  
also calculate an average of all readings taken since the  
MIN MAX mode was activated. To use MIN MAX mode,  
refer to the functions in Table 8.  
The high-accuracy 1 second response time has the full  
accuracy of the meter and is best for recording power  
supply drift, line voltage changes, or circuit performance  
while line voltage, temperature, load, or some other  
parameter is being changed.  
The true average value (AVG) displayed in the 100 ms  
and 1 s modes is the mathematical integral of all readings  
taken since you started recording. The average reading is  
useful for smoothing out unstable inputs, calculating  
power consumption, or estimating the percent of time a  
circuit is active.  
Response time is the length of time an input must stay at  
a new value to be recorded. A shorter response time  
captures shorter events, but with decreased accuracy.  
Changing the response time erases all recorded  
readings. Models 83 and 85 have 100 millisecond and  
1 second response times; Model 87 has 1 second,  
100 millisecond, and 250 µs (peak) response times. The  
250 µs response time is indicated by "1 ms" on the  
display.  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
MIN MAX Recording Mode  
Table 8. MIN MAX Functions  
MIN MAX Function  
Button  
M
Enter MIN MAX recording mode. The meter is locked in the range displayed before you  
entered MIN MAX mode. (Select the desired measurement function and range before  
entering MIN MAX.) The meter beeps each time a new minimum or maximum value is  
recorded.  
M
Scroll through minimum (MIN), maximum (MAX), and average (AVG) values.  
(While in MIN MAX mode)  
T
PEAK MIN MAX  
Model 87 only: Select 100 ms or 250 µs response time. (The 250 µs response time is  
indicated by "1 ms" on the display.) Stored values are erased. The present and AVG  
(average) values are not available when 250 µs is selected.  
I
Stop recording without erasing stored values. Press again to resume recording.  
Exit MIN MAX mode. Stored values are erased. The meter stays in the selected range.  
M
(hold for 1 second)  
Hold down M  
while turning the meter on  
Select 1 s high-accuracy response time. See text under "MIN MAX Recording Mode" for  
more explanation. MIN MAX readings for the frequency counter are recorded only in the  
high-accuracy mode.  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Touch Hold ® Mode  
In relative mode, the reading shown is always the  
difference between the present reading and the stored  
reference value. For example, if the stored reference  
value is 15.00 V and the present reading is 14.10 V, the  
display shows -0.90 V.  
WWarning  
The Touch Hold mode will not capture  
unstable or noisy readings. Do not use  
Touch Hold mode to determine that circuits  
are without power.  
On Model 87, the relative mode does not change the  
operation of the analog display.  
The Touch Hold mode captures the present reading on  
the display. When a new, stable reading is detected, the  
meter beeps and displays the new reading. To enter or  
exit Touch Hold mode, press I.  
Zoom Mode (Models 83 and 85)  
Selecting relative mode on a Model 83 or 85 meter  
causes the bar graph to enter Zoom mode. In zoom  
mode, the center of the graph represents zero and the  
sensitivity of the bar graph increases by a factor of 10.  
Measured values more negative than the stored  
reference light segments to the left of center; values more  
positive light segments to the right of center.  
Relative Mode  
Selecting relative mode ( C) causes the meter to  
zero the display and store the present reading as the  
reference for subsequent measurements. The meter is  
locked into the range selected when you pressed  
C. Press Cagain to exit this mode.  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
Uses for the Zoom Mode (Models 83 and 85)  
Maintenance  
The relative mode, combined with the increased  
sensitivity of the bar graph’s zoom mode, helps you make  
fast and accurate zero and peak adjustments.  
Repairs or servicing not covered in this manual should be  
performed only by qualified personnel as described in the  
80 Series III Service Manual.  
For zero adjustments, set the meter to the desired  
function, short the test leads together, press C;  
then connect the leads to the circuit under test. Adjust the  
circuit’s variable component until the display reads zero.  
Only the center segment on the Zoom bar graph is lit.  
General Maintenance  
Periodically wipe the case with a damp cloth and mild  
detergent. Do not use abrasives or solvents.  
Dirt or moisture in the terminals can affect readings and  
can falsely activate the Input Alert feature. Clean the  
terminals as follows:  
For peak adjustments, set the meter to the desired  
function, connect the leads to the circuit under test; then  
press C. The display reads zero. As you adjust for  
a positive or negative peak, the bar graph length  
increases to the right or left of zero. If an overange  
symbol lights  
1. Turn the meter off and remove all test leads.  
2. Shake out any dirt that may be in the terminals.  
3. Soak a new swab with a cleaning and oiling agent  
(such as WD-40). Work the swab around in each  
terminal. The oiling agent insulates the terminals  
from moisture-related activation of the Input Alert  
feature.  
(Û  
®
), press C twice to set a new  
reference; then continue with your adjustment.  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Testing the Fuses  
TRUE RMS MULTIMETER  
87  
Good F2 fuse: 00.0 to  
00.5 Ω  
Replace fuse: OL  
Before measuring current, test the appropriate fuse as  
shown in Figure 9. If the tests give readings other than  
those shown, have the meter serviced.  
MIN MAX  
RANGE  
REL  
HOLD  
Hz  
H
PEAK MIN MAX  
mV  
mA  
A
V
µA  
V
OFF  
WWarning  
Touch top half  
of input contacts  
A
mA µA  
COM  
V
!
400mA MAX  
CAT II  
FUSED  
10A MAX  
FUSED  
1000V MA
To avoid electrical shock or personal injury,  
remove the test leads and any input signals  
before replacing the battery or fuses. To  
prevent damage or injury, install ONLY  
specified replacement fuses with the  
amperage, voltage, and speed ratings shown  
in Table 9.  
!
TRUE RMS MULTIMETER  
87  
MIN MAX  
RANGE  
REL  
HOLD  
Hz  
H
PEAK MIN MAX  
mV  
mA  
A
V
Good F1 fuse: 0.995 kto  
1.005 kΩ  
µA  
V
OFF  
A
mA µA  
COM  
V
Replace fuse: OL  
!
400mA MAX  
CAT II  
FUSED  
10
F
1000V MA
!
iy5f.eps  
Figure 9. Testing the Current Fuses  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Maintenance  
Replacing the Battery  
Replacing the Fuses  
Replace the battery with a 9 V battery (NEDA A1604,  
6F22, or 006P).  
Referring to Figure 10, examine or replace the meter’s  
fuses as follows:  
1. Turn the rotary switch to OFF and remove the test  
leads from the terminals.  
WWarning  
To avoid false readings, which could lead to  
possible electric shock or personal injury,  
replace the battery as soon as the battery  
indicator (B) appears.  
2. Remove the battery door by using a standard-blade  
screwdriver to turn the battery door screws one-  
quarter turn counterclockwise.  
Replace the battery as follows (refer to Figure 10):  
3. Remove the three Phillips-head screws from the  
case bottom and turn the case over.  
1. Turn the rotary switch to OFF and remove the test  
leads from the terminals.  
4. Gently lift the input terminal-end of the top case to  
separate the two halves of the case.  
2. Remove the battery door by using a standard-blade  
screwdriver to turn the battery door screws one-  
quarter turn counterclockwise.  
5. Remove the fuse by gently prying one end loose,  
then sliding the fuse out of its bracket.  
3. Replace the battery and the battery door. Secure the  
door by turning the screws one-quarter turn  
clockwise.  
6. Install ONLY specified replacement fuses with the  
amperage, voltage, and speed ratings shown in  
Table 9.  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
6. Verify that the rotary switch and the circuit board  
switch are in the OFF position.  
Service and Parts  
If the meter fails, check the battery and fuses. Review this  
manual to verify proper use of the meter.  
7. Replace the case top, ensuring that the gasket is  
properly seated and case snaps together above the  
LCD (item A).  
Replacement parts and accessories are shown in Tables  
9 and 10 and Figure 11.  
8. Reinstall the three screws and the battery door.  
Secure the door by turning the screws one-quarter  
turn clockwise.  
To contact Fluke, call one of the following telephone  
numbers:  
USA: 1-888-99-FLUKE (1-888-993-5853)  
Canada: 1-800-36-FLUKE (1-800-363-5853)  
Europe: +31 402-678-200  
Japan: +81-3-3434-0181  
Singapore: +65-738-5655  
Anywhere in the world: +1-425-356-5500  
Or, visit Fluke’s Web site at www.fluke.com.  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service and Parts  
F1  
F2  
1
iy12f.eps  
Figure 10. Battery and Fuse Replacement  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Table 9. Replacement Parts  
Description  
Fluke Part  
or Model  
Number  
Item  
Quantity  
BT1  
Battery, 9 V  
614487  
943121  
803293  
832246  
824466  
831933  
1611720  
1611712  
688168  
688645  
1
1
1
3
2
1
1
1
1
F1 W  
F2 W  
H1  
MP1  
MP2  
TM1  
TM2  
TM3  
TM4  
Fuse, 0.440 A, 1000 V, FAST  
Fuse, 11 A, 1000 V, FAST  
Screw, Case  
Foot, Non-Skid  
O-Ring, Input Receptacle  
CD-ROM (contains Users Manual)  
Getting Started Manual  
Quick Reference Guide, Fluke 80 Series III  
Service Manual  
Optional  
WTo ensure safety, use exact replacement only.  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service and Parts  
TL75  
Test Lead Set  
MP85  
T24 Test  
Lead Set  
S1  
AC20 Alligator  
Clip (Black)  
TP1, TP4 Probes  
87/E Test Lead Set  
F2  
AC70A  
Alligator Clips  
F1  
MP2  
TM1  
C81Y  
MP86  
TM2  
H1  
MP1  
MP92  
BT1  
TM3  
H5, 6  
iy11f.eps  
Figure 11. Replaceable Parts  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Table 10. Accessories*  
Description  
Fluke Part  
Number  
Item  
Quantity  
TL20  
AC70A  
TL75  
TL24  
TP1  
Industrial Test Lead Set (Optional)  
Alligator Clips for use with TL75 test lead set  
Test Lead Set  
Test Lead Set, Heat-Resistant Silicone  
Test Probes, Flat Blade, Slim Reach  
Test Probes, 4 mm diameter, Slim Reach  
Safety Grip, Wide-Jaw Alligator Clips  
Holster, Yellow  
TL20  
AC70A  
TL75  
TL24  
TP1  
1
1
TP4  
TP4  
AC20  
C81Y  
C81G  
C25  
AC20  
C81Y  
C81G  
C25  
1
Holster, Gray (Optional)  
Carrying Case, Soft (Optional)  
* Fluke accessories are available from your authorized Fluke distributor.  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Specifications  
Specifications  
Maximum Voltage between any Terminal and Earth Ground: 1000 V rms  
WFuse Protection for mA or µA inputs: 44/100 A, 1000 V FAST Fuse  
WFuse Protection for A input: 11 A, 1000 V FAST Fuse  
Display: Digital: 4000 counts updates 4/sec; (Model 87 also has 19,999 counts in 4½-digit mode, updates 1/sec.). Analog: updates  
40/sec. Frequency: 19,999 counts, updates 3/sec at >10 Hz. Model 87: 4 x 32 segments (equivalent to 128); Models 83, 85: 43 segments.  
Temperature: Operating: -20°C to +55°C; Storage: -40°C to +60°C  
Altitude: Operating: 2000 m; Storage: 10,000 m  
Temperature Coefficient: 0.05 x (specified accuracy)/ °C (<18°C or >28°C)  
Electromagnetic Compatibility: In an RF field of 3 V/m total accuracy = specified accuracy except: Models 85,87: Total Accuracy =  
Specified Accuracy + 0.4% of range above 800 MHz (µADC only). (mVAC and µAAC unspecified). Model 83: Total Accuracy = Specified  
Accuracy + 5% of range above 300 MHz (µADC only). (VDC unspecified).  
Relative Humidity: 0% to 90% (0°C to 35°C); 0% to 70% (35°C to 55°C)  
Battery Type: 9 V zinc, NEDA 1604 or 6F22 or 006P  
Battery Life: 400 hrs typical with alkaline (with backlight off)  
Shock Vibration: Per MIL-T-28800 for a Class 2 instrument  
Size (HxWxL): 1.25 in x 3.41 in x 7.35 in (3.1 cm x 8.6 cm x 18.6 cm)  
Size with Holster and Flex-Stand: 2.06 in x 3.86 in x 7.93 in (5.2 cm x 9.8 cm x 20.1 cm)  
Weight: 12.5 oz (355 g)  
Weight with Holster and Flex-Stand: 22.0 oz (624 g)  
Safety: Complies with ANSI/ISA S82.01-1994, CSA 22.2 No. 1010.1:1992 to 1000 V Overvoltage Category III. UL listed to UL3111-1.  
Licensed by TÜV to EN61010-1.  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Table 11. Models 85 and 87 AC Voltage Function Specifications  
1
Function  
Range  
Resolution  
Accuracy  
2
50 Hz - 60 Hz  
45 Hz - 1 kHz  
1 kHz - 5 kHz  
5 kHz - 20 kHz  
400.0 mV  
4.000 V  
40.00 V  
400.0 V  
1000 V  
0.1 mV  
0.001 V  
0.01 V  
0.1 V  
±(0.7% + 4)  
±(0.7% + 2)  
±(0.7% + 2)  
±(0.7% + 2)  
±(0.7% + 2)  
±(1.0% + 4)  
±(1.0% + 4)  
±(1.0% + 4)  
±(1.0% + 4)  
±(2.0% + 4)  
±(2.0% + 4)  
±(2.0% + 4)  
±(2.0% + 20)  
±(2.0% + 20)  
±(2.0% + 20)  
unspecified  
unspecified  
K 3  
4
±(2.0% + 4)  
5
1 V  
±(1.0% + 4)  
unspecified  
1. Accuracy is given as ±([% of reading] + [number of least significant digits]) at 18°C to 28°C, with relative humidity up to 90%, for a  
period of one year after calibration. For Model 87 in the 4 ½-digit mode, multiply the number of least significant digits (counts) by 10.  
AC conversions are ac-coupled and valid from 5% to 100% of range. Models 85 and 87 are true rms responding. AC crest factor can  
be up to 3 at full scale, 6 at half scale. For non-sinusoidal wave forms add -(2% Rdg + 2% full scale) typical, for a crest factor up to 3.  
2. Below 10% of range, add 6 counts.  
3. Models 85 and 87 are true rms responding meters. When the input leads are shorted together in the ac functions, the meters display  
a reading (typically <25 counts) that is caused by internal amplifier noise. The accuracy on Models 85 and 87 is not significantly  
affected by this internal offset when measuring inputs that are within 5% to 100% of the selected range. When the rms value of the  
two values (5% of range and internal offset) is calculated, the effect is minimal as shown in the following example where 20.0 = 5% of  
400 mV range, and 2.5 is the internal offset: RMS = SQRT[(20.0)2 + (2.5)2] = 20.16. If you use the REL function to zero the display  
when using the ac functions, a constant error that is equal to the internal offset will result.  
4. Frequency range: 1 kHz to 2.5 kHz.  
5. Below 10% of range, add 16 counts.  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Specifications  
Table 12. Model 83 AC Voltage Function Specifications  
1
Function  
Range  
Resolution  
Accuracy  
50 Hz - 60 Hz  
45 Hz - 1 kHz  
1 kHz - 5 kHz  
400.0 mV  
4.000 V  
40.00 V  
400.0 V  
1000 V  
0.1 mV  
0.001 V  
0.01 V  
0.1 V  
±(0.5% + 4)  
±(0.5% + 2)  
±(0.5% + 2)  
±(0.5% + 2)  
±(0.5% + 2)  
±(1.0% + 4)  
±(1.0% + 4)  
±(1.0% + 4)  
±(1.0% + 4)  
±(1.0% + 4)  
±(2.0% + 4)  
±(2.0% + 4)  
±(2.0% + 4)  
K2  
3
±(2.0% + 4)  
1 V  
unspecified  
1. See the first sentence in Table 11 for a complete explanation of accuracy.  
2. Below a reading of 200 counts, add 10 counts.  
3. Frequency range: 1 kHz to 2.5 kHz.  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Table 13. DC Voltage, Resistance, and Conductance Function Specifications  
1
Accuracy  
Function  
Range  
4.000 V  
Resolution  
Model 83  
Model 85  
Model 87  
0.001 V  
0.01 V  
0.1 V  
1 V  
±(0.1% + 1)  
±(0.1% + 1)  
±(0.1% + 1)  
±(0.1% + 1)  
±(0.08% + 1)  
±(0.08% + 1)  
±(0.08% + 1)  
±(0.08% + 1)  
±(0.05% + 1)  
±(0.05% + 1)  
±(0.05% + 1)  
±(0.05% + 1)  
L
40.00 V  
400.0 V  
1000 V  
400.0 mV  
0.1 mV  
±(0.3% + 1)  
±(0.1% + 1)  
±(0.1% + 1)  
F
mV  
2
2
2
400.0 Ω  
0.1 Ω  
±(0.4% + 2)  
±(0.4% + 1)  
±(0.4% + 1)  
±(0.7% + 1)  
±(0.7% + 1)  
±(1.0% + 3)  
±(1.0% + 10)  
±(0.2% + 2)  
±(0.2% + 2)  
e
4.000 kΩ  
40.00 kΩ  
400.0 kΩ  
4.000 MΩ  
40.00 MΩ  
40.00 nS  
0.001 kΩ  
0.01 kΩ  
0.1 kΩ  
0.001 MΩ  
0.01 MΩ  
0.01 nS  
±(0.2% + 1)  
±(0.2% + 1)  
±(0.6% + 1)  
±(0.6% + 1)  
±(1.0% + 3)  
±(1.0% + 10)  
±(0.2% + 1)  
±(0.2% + 1)  
±(0.6% + 1)  
±(0.6% + 1)  
±(1.0% + 3)  
±(1.0% + 10)  
nS  
1. See the first sentence in Table 11 for a complete explanation of accuracy.  
2. When using the REL function to compensate for offsets.  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Specifications  
Table 14. Current Function Specifications  
Accuracy  
1
2
3, 4  
3, 4  
Function  
Range  
Resolution  
Model 83  
Model 85  
Model 87  
Burden Voltage  
(typical)  
6
6
40.00 mA  
400.0 mA  
4000 mA  
0.01 mA  
0.1 mA  
1 mA  
±(1.2% + 2)  
±(1.2% + 2)  
±(1.2% + 2)  
±(1.2% + 2)  
±(1.0% + 2)  
±(1.0% + 2)  
±(1.0% + 2)  
±(1.0% + 2)  
±(1.0% + 2)  
±(1.0% + 2)  
±(1.0% + 2)  
±(1.0% + 2)  
1.8 mV/mA  
1.8 mV/mA  
0.03 V/A  
mA  
\
(45 Hz to 2 kHz)  
6
6
6
6
6
6
5
10.00 A  
0.01 A  
0.03 V/A  
mA  
[
±(0.4% + 4)  
±(0.4% + 2)  
±(0.4% + 4)  
±(0.4% + 2)  
±(0.2% + 4)  
±(0.2% + 2)  
±(0.2% + 4)  
±(0.2% + 2)  
±(0.2% + 4)  
±(0.2% + 2)  
±(0.2% + 4)  
±(0.2% + 2)  
40.00 mA  
400.0 mA  
4000 mA  
0.01 mA  
0.1 mA  
1 mA  
1.8 mV/mA  
1.8 mV/mA  
0.03 V/A  
5
10.00 A  
0.01 A  
0.03 V/A  
1. See the first sentence in Table 11 for a complete explanation of accuracy.  
2. AC conversion for Model 83 is ac coupled and calibrated to the rms value of a sinewave input.  
3. AC conversions for Models 85 and 87 are ac coupled, true rms responding, and valid from 5% to 100% of range.  
4. See note 3 in Table 11.  
5.  
10 A continuous; 20 A for 30 seconds maximum; >10 A: unspecified.  
6. Below a reading of 200 counts, add 10 counts.  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Table 14. Current Function Specifications (continued)  
1
Accuracy  
2
3, 4  
3, 4  
Function  
Range  
Resolution  
Model 83  
Model 85  
Model 87  
Burden  
Voltage  
(typical)  
µA B  
(45 Hz to 2 kHz)  
5
5
5
400.0 µA  
4000 µA  
0.1 µA  
1 µA  
±(1.2% + 2)  
±(1.2% + 2)  
±(1.0% + 2)  
±(1.0% + 2)  
±(1.0% + 2)  
±(1.0% + 2)  
100 µV/µA  
100 µV/µA  
5
µAF  
400.0 µA  
4000 µA  
0.1 µA  
1 µA  
±(0.4% + 4)  
±(0.4% + 2)  
±(0.2% + 4)  
±(0.2% + 2)  
±(0.2% + 4)  
±(0.2% + 2)  
100 µV/µA  
100 µV/µA  
1. See the first sentence in Table 11 for a complete explanation of accuracy.  
2. AC conversion for Model 83 is ac coupled and calibrated to the rms value of a sinewave input.  
3. AC conversions for Models 85 and 87 are ac coupled, true rms responding, and valid from 5% to 100% of range.  
4. See note 3 in Table 11.  
5. Below a reading of 200 counts, add 10 counts.  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Table 15. Capacitance and Diode Function Specifications  
1
Function  
Range  
Resolution  
Accuracy  
5.00 nF  
0.01 nF  
±(1% + 3)  
±(1% + 3)  
±(1% + 3)  
E
0.0500 µF  
0.500 µF  
5.00 µF  
0.0001 µF  
0.001 µF  
0.01 µF  
±(1.9% + 3)  
3.000 V  
0.001 V  
±(2% + 1)  
G
1. With a film capacitor or better, using Relative mode to zero residual. See the first sentence in Table 11 for a complete explanation of  
accuracy.  
Table 16. Frequency Counter Specifications  
1
Function  
Frequency  
Range  
199.99  
Resolution  
Accuracy  
0.01 Hz  
0.1 Hz  
±(0.005% + 1)  
±(0.005% + 1)  
±(0.005% + 1)  
±(0.005% + 1)  
unspecified  
(0.5 Hz to 200 kHz,  
1999.9  
pulse width >2 µs)  
19.999 kHz  
199.99 kHz  
>200 kHz  
0.001 kHz  
0.01 kHz  
0.1 kHz  
1. See the first sentence in Table 11 for a complete explanation of accuracy.  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Table 17. Frequency Counter Sensitivity and Trigger Levels  
Minimum Sensitivity (RMS Sinewave) Approximate Trigger Level  
1
Input Range  
5 Hz - 20 kHz  
0.5 Hz - 200 kHz  
(DC Voltage Function)  
400 mV dc  
400 mV dc  
4 V  
70 mV (to 400 Hz)  
70 mV (to 400 Hz)  
150 mV  
40 mV  
150 mV  
0.3 V  
3 V  
0.7 V  
1.7 V  
4 V  
40 V  
7 V (140 kHz)  
70 V (14.0 kHz)  
700 V (1.4 kHz)  
400 V  
30 V  
40 V  
400 V  
1000 V  
300 V  
Duty Cycle Range  
Accuracy  
0.0 to 99.9%  
Within ±(0.05% per kHz + 0.1%) of full scale for a 5 V logic family input on the 4 V dc range.  
Within ±((0.06 x Voltage Range/Input Voltage) x 100%) of full scale for sine wave inputs on ac voltage ranges.  
1. Maximum input for specified accuracy = 10X Range or 1000 V.  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Specifications  
Table 18. Electrical Characteristics of the Terminals  
Input  
Common Mode Rejection  
Ratio  
Overload  
Impedance  
(nominal)  
1
Function  
Protection  
(1 kunbalance)  
Normal Mode Rejection  
>60 dB at 50 Hz or 60 Hz  
>60 dB at 50 Hz or 60 Hz  
L
1000 V rms  
1000 V rms  
10 M<100 pF  
10 M<100 pF  
>120 dB at dc, 50 Hz or 60 Hz  
>120 dB at dc, 50 Hz or 60 Hz  
F
mV  
K
1000 V rms  
10 M<100 pF  
>60 dB, dc to 60 Hz  
(ac-coupled)  
Full Scale Voltage  
Typical Short Circuit Current  
Open Circuit  
Test Voltage  
To 4.0 MΩ  
40 Mor nS  
400 Ω  
4 k  
40 k  
400 k  
4 M  
40 M  
e
1000 V rms  
1000 V rms  
<1.3 V dc  
<3.9 V dc  
<450 mV dc  
<1.3 V dc  
200 µA 80 µA  
12 µA  
1.4 µA  
0.2 µA 0.2 µA  
G
3.000 V dc  
0.6 mA typical  
6
1. 10 V Hz max  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
80 Series III  
Users Manual  
Table 19. MIN MAX Recording Specifications  
Accuracy  
Model  
Nominal Response  
83  
100 ms to 80%  
1 s  
Specified accuracy ±12 counts for changes >200 ms in duration (±40 counts in ac with beeper on)  
Same as specified accuracy for changes >2 seconds in duration (±40 counts in ac with beeper on)  
85, 87  
100 ms to 80%  
(DC functions)  
Specified accuracy ±12 counts for changes >200 ms in duration  
Specified accuracy ±40 counts for changes >350 ms and inputs >25% of range  
Same as specified accuracy for changes >2 seconds in duration  
120 ms to 80%  
(AC functions)  
1 s  
250 µs  
(Model 87 only)  
Specified accuracy ±100 counts for changes >250 µs in duration  
(± 250 digits typical for mV, 400 µA dc, 40 mA dc, 4000 mA dc)  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 

Echo Trimmer SRM230 User Manual
Eiki Projector 3990BDA User Manual
Electro Voice Microphone N D308 User Manual
Extron electronic Network Card CPM200 User Manual
Focal Car Speaker 30 A1 User Manual
FujiFilm Film Camera RVP100 User Manual
Furuno Thermometer TI 20 User Manual
Generac Portable Generator 005938 0 User Manual
Genesis ICE Speaker Genesis 52 User Manual
Global Machinery Company Saw CDB305MS User Manual