Agilent Technologies Thermometer 1141A User Manual

User and Service Guide  
Publication number 01141-97002  
July 2004  
For Safety and Regulatory information, see the pages at the end  
of this book.  
©
Copyright Agilent Technologies 2000-2004  
All Rights Reserved  
Agilent Technologies 1141A  
Differential Probe and 1142A  
Probe Control and Power  
Module  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Accessories Supplied  
6
Recommended Test Equipment  
9
Calibration Test Record 39  
Probe Adjustment 40  
Attenuator Adapter Adjustment 49  
3 Service  
Introduction 54  
Performance Specifications and Characteristics 55  
General Characteristics 58  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Differential Probe 60  
Probe Control and Power Module Troubleshooting 64  
Replaceable Parts 71  
Parts List 71  
Ordering Information 71  
Direct Mail Order System 71  
Manufacturers’ Codes 72  
Exploded View 73  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
1
Operating the Probe  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating the Probe  
Accessories Supplied  
Introduction  
This chapter shows you how to connect and operate the 1141A  
Differential Probe and 1142A Probe Control and Power Module as a  
differential probe system.  
Accessories Supplied  
The following items are supplied as part of the 1141A/1142A probe  
system. Item numbers refer to the numbers in Figure 1-1 on page -7  
and Figure 1-2 on page -8. Those without item numbers are supplied  
but not shown in figures. See the Replaceable Parts List for parts not  
listed below.  
Item  
1
Description  
Qty  
1
Part Number  
Differential Probe  
2
10x Attenuator Adapter  
100x Attenuator  
1
5063-2144  
5063-2145  
5063-2146  
5959-9334  
1400-1422  
5061-6162  
01141-68702  
01141-66504  
8710-1961  
1251-5943  
1142A  
3
1
4
ac Coupling Adapter  
Two-inch Extension Leads (package 5)  
Mini Grabbers  
1
5
1
6
2
7
Five-inch Ground Lead  
Shielded Signal Lead  
Test Board  
1
8
1
9
1
10  
11  
Flat-blade Alignment Tool  
Circuit Connection Posts (strip of 20)  
Probe Control and Power Module  
Power Cord  
1
1
1
1
see parts list  
Carrying case  
User and Service Manual  
1
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating the Probe  
Accessories Supplied  
Figure 1-1  
1141A Differential Probe and Accessories  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating the Probe  
Accessories Available  
Figure 1-2  
1141A Miscellaneous Accessories  
Figure 1-3  
1142A Probe Control and Power Module  
Accessories Available  
The following accessories can be ordered.  
5959-9335 Long Extension Lead (5.5 inch/14 cm), package of 5  
5090-4833 Mini grabber for SMT, package of 20  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating the Probe  
To inspect the probe  
Inspect the shipping container for damage. If the shipping container or cushioning  
material is damaged, it should be kept until the contents of the shipment have been  
checked for completeness and the instrument had been checked mechanically and  
electrically. Accessories supplied with the instrument are listed in see “Accessories  
Supplied” on page 6 of this manual.  
If the contents are incomplete, if there is mechanical damage or defect, or if the  
instrument does not pass calibration tests, notify the nearest Agilent Technologies office.  
If the shipping container is damaged, or the cushioning materials show sign of stress,  
notify the carrier as well as the nearest Agilent Technologies office. Keep the shipping  
materials for the carrier’s inspection. The office will arrange for repair or replacement  
at Agilent Technologies’ option without waiting for a claim settlement.  
Using the probe with other instruments  
The 1141A/1142A probe system can be used with other instruments as well as  
oscilloscopes. You can use it with a spectrum analyzer or frequency counter, or any  
instrument with an input that can be terminated with 50 .  
If you are going to use the probe system with an instrument other than an oscilloscope,  
you may need to set up the probe with an oscilloscope first. This will allow you to select  
coupling and reject modes, and set offset, so the output of the probe is compatible with  
signal requirements of the other instrument.  
Recommended Test Equipment  
The following table is a list of the test equipment required to test calibration, make  
adjustments, and troubleshoot this instrument. The table indicates the critical  
specificationsofthetest equipmentandfor whichproceduretheequipmentisnecessary.  
Equipment other than the recommended model may be used if it satisfies the critical  
specifications listed in the table.  
Recommended Test Equipment  
Recommended  
Agilent  
Equipment Required Critical Specifications  
Model/Part  
Use  
P
Signal Generator  
Function Generator  
Oscilloscope  
1 to 200 MHz, 400 m V  
rms  
8648A  
2.5 to 500 kHz, 600 m V to 16 V  
33120A  
A
p-p  
p-p  
>400 MHz bandwidth, 1 mV/div sensitivity 54830B  
100 mV to 7 Vdc E3632A  
P,A  
P
dc source  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Operating the Probe  
Recommended Test Equipment  
Recommended  
Agilent  
Equipment Required Critical Specifications  
Model/Part  
Use  
A
Test Board  
Load  
No substitute (supplied accessory)  
01141-66504  
BNC Feedthrough, 50 Ω  
Pasternack  
P
Enterprises PE6008-  
50 or Huber+Suhner  
22543742  
Cables (2)  
Cable  
BNC, 50 36-inch  
10503A  
P,A  
A
BNC, 50 9-inch  
10502A  
Cable  
Type N (m) 24-inch  
11500B  
P
Adapter  
Type N (m) to BNC (f)  
Type N (f) to BNC (m)  
BNC (f) to dual banana (m)  
Small flat blade (supplied accessory)  
1250-0780  
1250-0077  
1251-2277  
A
Adapter  
P
Adapters (2)  
Alignment tool  
P
8710-1961/  
A
Sprague-Goodman  
part number GTT-5G  
P=Calibration Tests, A=Adjustments  
Setting up the probe  
The following paragraphs cover system preparation and initial adjustments.  
Power Requirements  
The 1141A/1142A probe system (specifically the 1142A) requires a power source of  
either 90 to 132/198 to 264 Vac, 47 to 440 Hz, 25 VA maximum.  
C A U T I O N  
Before connecting power to this instrument, be sure the line voltage switch on the rear  
panel of the instrument is set properly.  
Line Voltage Selection  
Before applying power, verify the setting of the LINE SELECT switch on the rear panel  
of the 1142A. The slide switch can be set to either 115 or 230 V.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating the Probe  
Recommended Test Equipment  
W A R N I N G  
Before connecting this instrument, the protective earth terminal of the instrument must  
be connected to the protective conductor of the (Mains) power cord. The Mains plug  
must be inserted in a socket outlet provided with a protective each contact. The  
protective action must be negated by the use on an extension cord (power cable)  
without a protective conductor (grounding). Grounding one conductor of a two-  
conductor outlet does not provide an instrument ground.  
This instrument is provided with a three-wire power cable. When connected to an  
appropriate ac power outlet, this cable grounds the instrument cabinet. The type of  
power cable plug shipped with the instrument depends on the country of destination.  
The 1142A Power Control and Power Module does not have a power switch. A power  
switch is not required because of the low mains power requirement.  
Figure 1-4  
1142A Rear Panel  
Procedure  
1 Use the power cord to connect the 1142A to the ac mains.  
2 Connect the 1141A probe cable power connector to the PROBE connector on  
the rear panel of the 1142A power module.  
3 Connect the output of the probe to the input of the oscilloscope.  
4 Set the input impedance of the oscilloscope to 50 .  
If the oscilloscope does not have a selectable 50 input impedance, connect a 50 BNC  
feedthrough termination between the probe output and the input of the oscilloscope.  
5 If making an initial equipment setup, continue with the initial adjustment in  
the following section.  
Initial Adjustment  
For a given combination of 1141A Differential Probe and 1142A Probe Control and Power  
Module, you may want to adjust the Offset Null and DC Reject Gain. Typically, you need  
to make these adjustments only once, before the probe is first used. You can make them  
any time to optimize the system. These adjustments do not affect the specifications of  
the probe system.  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating the Probe  
Recommended Test Equipment  
Offset null zeroes the dc level at the output of the probe. The range of adjustment is  
about 4 mV.  
DC Reject Gain adjusts the gain of the dc reject circuit to accurately null the dc  
component of an input signal. The range of adjustment is about 0.5%.  
Figure 1-5  
1142A Front Panel  
Equipment Needed  
The following equipment is necessary for initial adjustment.  
• 5 V power supply  
• DVM that can measure 25 µV  
• 50 BNC feedthrough terminator  
• 01141-66504 test board  
Equipment Setup  
Use the following procedure to setup the differential probe system for initial adjustment.  
C A U T I O N  
Do not exceed 7 V when using the test board for this procedure. If the voltage is too  
high, it will cause excessive power dissipation in the 50 termination on the test board.  
1 Use the probe setup procedure to set up the probe system.  
2 Connect a 50 BNC feedthrough terminator to the output of the probe.  
3 Disconnect all accessories from the input of the probe.  
4 Connect the DVM to measure the dc output of the probe at the 50 load.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating the Probe  
Recommended Test Equipment  
5 Set up the 1142A:  
a
Set the Local/Remote push button to Local.  
b
Under DC Couple, press Zero offset.  
6 Set the power supply output to 5 V.  
7 Arrange a connection between the power supply and the test board. The  
negative terminal of the supply should connect to the shield of the test board  
BNC.  
If your power supply has standard binding posts, you can connect a banana-to-BNC  
adapter to the supply and connect a BNC cable between the supply and the test board.  
Adjustment  
Warm up the 1141A for 30 minutes before making adjustments.  
1 With the 1141A probe inputs unconnected, adjust Offset Null on the 1142A for  
a minimum reading on the DVM.  
The voltage swing of the adjustment is approximately 4 mV.  
2 On the 1142A, under DC Reject, press 5.0 Hz.  
3 Read and record the reading on the DVM, _________ mV.  
4 Connect the probe to the test board in the position shown below  
(signal to + input).  
Figure 1-6  
Signal to + input  
5 AftertheDVMreadingstabilizes, adjustDCRejectGaintothereadingrecorded  
in step 3.  
With a 5 V supply, the voltage swing is approximately 12.5 mV. With a lower supply, the  
voltage swings proportionally less.  
Using the Accessories  
The 1141A Differential Probe and accessories are designed to provide a variety of ways  
to connect to circuitry and make measurements. In the descriptions, any method used  
to connect to the probe signal inputs also applied to the adapters. The figure below  
shows, in a general way, the use of accessories.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating the Probe  
Recommended Test Equipment  
Figure 1-7  
Basic Accessory Connections  
Probe Tips  
Probe tips fit into the receptacles in the probe and are held in place with probe tip caps.  
If necessary, you can solder the probe tips into a circuit or wires can be soldered to the  
tips. If you solder to the probe tips, be careful not to melt the plastic probe tip caps.  
N o t e  
Because of the close tolerances between the probe tip caps and probe tips, it will be  
difficult to separate the probe tips and caps once the probe tips have been soldered.  
Ground Leads  
The circular end of the ground lead fits over the screw on the top side of the probe.  
Extension Leads  
The extension leads provide a flexible connection between circuitry and the probe.  
To provide a male connection to other circuitry, connect the extension lead over the  
probe tips.  
Toprovideafemaleconnection, removetheprobetipcapsandprobetipsandconnect  
the extension leads to the probe.  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating the Probe  
Recommended Test Equipment  
N o t e  
Use extension leads and similar connection accessories carefully. Extension leads  
compromise the high-frequency specifications of the probe. CMRR is particularly  
sensitive to unbalanced input parameters.  
To prevent pickup of stray fields when you use extension lead, either the ones supplied  
with the 1141A or others, dress them carefully as follows:  
Connect the leads at right angles to the circuitry under test.  
Keep the leads as parallel as possible before they connect to the probe.  
Mini-Grabbers  
Mini-grabbers can be attached to the probe or adapter through the extension leads.  
1 Remove the probe tip caps and tips.  
2 Attach the extension leads to the probe or adapter.  
3 Attach the mini-grabbers to the extension leads.  
Circuit Connector Posts  
These 0.025-inch square posts can be used to connect either directly to the probe or to  
the extension leads.  
Solder the posts directly into your circuitry or use them to make extension leads that  
plug into the inputs of the probe or adapters.  
Shielded Signal Leads  
The shielded signal leads allow connection to points in a circuit that are up to 10 inches  
apart. The leads are shielded to within approximately 1/2 inch of the end of the lead so  
they minimize pick-up due to stray fields from adjacent circuitry.  
1 Connect the end with the ground connector to the probe pins and ground of  
the differential probe or adapter.  
2 Connect the free ends of the leads to 0.025-inch square or 0.030-inch round pins  
in your circuitry or to the mini-grabbers.  
N o t e  
Each lead has an input capacitance of approximately 15 pF. This capacitance may limit  
the bandwidth of your measurement (depending on the impedance of the circuit). Also,  
CMRR may be affected because of slight differences between the input capacitance of  
the two leads. CMRR is also affected by differences in impedance between the two  
measurements points.  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating the Probe  
Recommended Test Equipment  
Adapters  
There are three adapters for use with the differential probe. Two adapters are  
attenuators, a 10x and a100x. The other is an ac adapter for blocking dc from the probe  
input.  
The adapters are installed on the probe after the probe tip caps and probe tips have been  
removed. The adapter fastens to the probe using a thumb wheel located on the underside  
of the adapter. the figure below shows a good way to hold the probe while attaching the  
adapter.  
1 Remove the probe tip caps and probe tips, from the probe.  
2 Fit the adapter over the end of the probe and rotate the thumb wheel with your  
finger until the adapter fits snugly.  
A snug fit is important because the ground is maintained through the thumb wheel screw.  
A loosely attached adapter compromises the mechanical and electrical integrity of the  
combination.  
Figure 1-8  
Attaching the Adapters  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating the Probe  
Recommended Test Equipment  
Adapter Combinations  
The figure below shows the allowed adapter and probe connections. There are two  
specific combinations that should not be used.  
Do not attach the ac adapter between an attenuator adapter and the probe.  
An attenuator adapter must be terminated by the input resistance of the probe. The  
ac adapter isolates the probe input resistance.  
Do not cascade two attenuator adapters.  
The attenuator adapters are designed to be terminated by the 1 Mresistance of the  
probe. The input resistance of the attenuator adapter is 9 Mfor the 10x adapter and  
10 Mfor the 100x adapter.  
Figure 1-9  
Allowed Adapter Connections  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating the Probe  
Recommended Test Equipment  
Connector Compatibility  
The following are general connector characteristics for the probe, adapters, and  
accessories.  
The female connectors on the probe, adapters, and other accessories are designed to  
mate with 0.030-inch round or 0.0250-inch square pins.  
The probe, adapter, and extension lead pins are 0.030-inch round.  
The strip of circuit connection posts provided as an accessory has 0.025-inch square  
pins.  
The mini-grabber has a 0.25-inch square pin.  
The ground connection at the end of the probe and adapters (where the adapters  
fasten) accepts an M3 metric screw.  
Test Board  
The primary use of the test board is to apply test and calibration signals to the input of  
the probe or adapters. Specific use of the test board is covered wherever it applies.  
Grounding  
Grounding is very important when probing circuitry. Improper grounding can increase  
the common mode signal level. This reduces the effectiveness of the differential probe.  
The mechanical connections at the input of the probe are ground for probe signals. The  
screw where the ground lead attaches (see figure 1-7) fastens to this ground. Also, the  
attenuator and ac adapter fasten to this ground through the screw connection and the  
ground is carried through each adapter to its front.  
Figure 1-10  
Probe Grounds  
Coupling Functions  
There are three methods for blocking or compensating for the dc component of a signal.  
Each has specific advantages.  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Operating the Probe  
Recommended Test Equipment  
Probe System Coupling Functions  
dc offset  
dc Reject  
ac Coupling  
Adapter  
dc Blocked  
20 V  
200 V  
500 V  
20 V  
200 V  
500 V  
20 V  
500 V  
500 V  
Probe alone  
Probe with 10x  
Probe with 100x  
Adjust offset to put  
signal on screen  
Select DC Reject  
low-frequency  
corner  
Attach ac Coupling  
adapter to  
differential probe  
Set-up needed  
1
1
Remote Control?  
Yes  
Yes  
No  
CMRR  
No  
No  
No  
Yes  
degradation?  
Low-frequency  
degradation?  
Yes  
Yes  
1
Isolated external dc reference and control signals are needed  
dc Reject  
dc Reject is the best method of eliminating the dc component of a signal when dc is not  
a factor in the measurement.  
The key characteristics are:  
The low-frequency component (from dc to the selected corner frequency) is  
automatically nulled by the dc reject circuitry.  
Probe CMRR specifications are not compromised as happens when the ac coupling  
adapter is uses.  
There is a selectable low frequency corner with -3dB points at 0.05 Hz, 0.5 Hz,  
or 5.0 Hz.  
Thevoltagerejectrangeis 20Vwiththeprobealone, 200Vwiththe10xattenuator,  
and 500 V with the 100x attenuator.  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating the Probe  
Recommended Test Equipment  
To use dc reject:  
1 Remove the ac adapter if it is installed.  
2 On the front panel of the 1142A, press Local.  
3 Under DC Reject on the front panel, press 5.0 Hz or 0.5 Hz individually, or  
5.0 Hz and 0.5 Hz simultaneously to get 0.05 Hz.  
Within the frequency and voltage characteristics noted elsewhere in this manual, low  
frequencies are nulled from the input signal.  
Offset  
Offset is the best method to use when the low-frequency corners associated with dc  
reject and the ac adapter interfere with the measurement.  
The key characteristics are:  
The user manually null the dc component with the offset adjustment.  
Offset is dc coupled so there is no low frequency roll-off.  
Probe CMRR specifications are not compromised as happens when the ac coupling  
adapter is used.  
Thevoltageoffsetrange is 20V with theprobealone, 200V with the10xattenuator,  
and 500 V with the 100x attenuator. (With the 100x attenuator, the offset range is  
restricted by the maximum input voltage rating rather than the operating range of  
the offset).  
To use offset:  
1 Remove the ac adapter if it is installed.  
2 On the front panel of the 1142A, press Local and Variable offset.  
3 Adjust the Coarse and Fine Variable Offset until the signal is displayed on the  
screen of the oscilloscope.  
ac Adapter  
The ac adapter must be used when the dc component of the signal exceeds the operating  
range of the dc reject or offset methods. The ac adapter block the dc and low frequency  
component of the input by forming a high pass filter with the input impedance of the  
probe or adapter.  
The key characteristics are:  
The ac adapter safely blocks 200 Vdc when attached directly to the probe or  
500 Vdc when attached to a 10x or 100x adapter.  
The probe and adapters have different input impedances, so they have different low  
frequency corners with the ac adapter. When the ac adapter is directly on the probe  
the -3dB corner is 15 Hz. When the ac adapter is on an attenuator the corner is 1.5 Hz.  
The low-frequency CMRR when using the ac adapter is not as good as when using the  
probe alone or the probe with a 10x or 100x adapter.  
C A U T I O N  
If you measure a node having a high dc potential, the blocking capacitors in the ac  
adapter will charge to that potential. After making such measurements, discharge the  
capacitors by grounding both inputs of the ac adapter. This will prevent damage by a  
high voltage discharge into sensitive circuitry when the next measurement is made.  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating the Probe  
Recommended Test Equipment  
To use ac coupling:  
1 Attach the ac coupling adapter to the input of the probe or the input of the  
attenuator adapter.  
2 On the 1142A, press Local and Zero offset.  
Remote operation  
For automatic test applications, the coupling and offset functions provided by the 1142A  
Probe Control and Power Module can be remotely controlled through a connector on the  
rear panel of the module. The connection is through a standard 9-pin female  
D-subminiature connector. This style is the same as that used on some personal  
computer monitor cables, which provides an economical way to connect the 1142A to  
the controller interface on an automatic test system.  
The following table gives the connections.  
Remote Input Connections  
Pin  
1
Function  
Pin  
6
Function  
Connector  
Function Select 1 (A1R)  
Digital common  
N.C.  
Function Select 0 (A0R)  
2
7
N.C.  
3
8
N.C.  
4
External offset common  
Shield  
9
External offset  
5
N O T E  
To minimize dc offset errors and potential noise coupling, electrically isolate all  
connections between the Remote Input connector and the controlling system.  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operating the Probe  
Recommended Test Equipment  
Function Select  
The easiest way to control the function select lines is contact closures between the lines  
and Digital common (pin 2) of the remote input connector. (TTL compatible control  
signals can be used; but to avoid problems with ground loops, they must be electrically  
isolated.) The following truth table shows the functions provided by the function select  
lines. For the Remote Inputs, “0” represents a closure and “1” represents an open circuit.  
Remote Functions Select Truth Table  
Remote Input  
Function  
A1R  
(Pin 1)  
A0R  
(Pin 6)  
0
0
1
1
0
1
0
1
0.05 Hz Reject  
0.5 Hz Reject  
5 Hz Reject  
DC Couple  
Variable Offset  
The remote variable offset can be used when the dc couple function is remotely selected.  
The offset voltage must be referenced to the External offset common (pin 4) of the  
remote input connector. It must be electrically isolated from the controlling system. The  
following table shows the offset range and remote offset requirements for probe and  
adapter combinations.  
Remote Offset Input Requirements  
Offset Range  
Remote  
Requirements  
Probe alone  
20 V  
200 V  
500 V  
10 V  
10 V  
2.5 V  
Probe with 10x adapter  
Probe with 100x adapter  
Cleaning Requirements  
If the instrument requires cleaning: (1) Remove power from the instrument. (2) Clean  
the external surfaces of the instrument with a soft cloth dampened with a mixture of  
mild detergent and water. (3) Make sure that the instrument is completely dry before  
reconnecting it to a power source.  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Operating the Probe  
Recommended Test Equipment  
Differential Amplifiers and CMRR  
The 1141A Differential Probe is a high-impedance differential amplifier. A characteristic  
of differential amplifiers is the ability to reject signals that are common to the two inputs.  
The common mode rejection ratio (CMRR) is the measurement of this ability. It is  
expressed as the ratio between the amplitudes of the common mode and differential  
signals which product equal outputs. For example, if a common mode signal of 1 V and  
differential signal of 1 mV both produce outputs of 1 mV, the CMRR is 1000:1.  
The ability to reject common mode signals is dependent on the balance designed into  
the differential amplifier. At higher frequencies it becomes harder to balance circuit  
parasitics and parameters of devices so CMRR degrades as frequency increases. Also,  
stray coupling increases with frequency and coupling may vary between the two  
differential paths.  
The CMRR of the 1141A Differential Probe is specified at the input of the probe and  
cannot be affected expect by adjustments in the probe. However, the way the probe is  
connected into the circuitry being tested can have a big influence in the overall result of  
the measurement, especially at high frequencies.  
The following things can affect the effective CMRR of a test setup:  
The connection to the circuit under test.  
The method used to connect the probe is important because it involves the symmetry  
of the differential input circuitry. For example, using different lengths of wire to  
connect the circuit to the two probe inputs unbalances the inductance and  
capacitance at the inputs. The effective CMRR will be reduced, especially at high  
frequencies. Additionally, coupling from adjacent circuitry will be less balanced.  
The impedance of the source.  
This is another instance where the symmetry of the differential circuit is important.  
The impedance of the source forms a network with the input impedance of the source  
forms a network with the input impedance of the connections and the probe. This  
network determines the frequency response for the measurement. If each side of the  
differential source has a different impedance, the frequency response of each side  
will be different and the unbalance is reflected in a reduced CMRR. Of course, lower  
source impedances have less effect on the frequency response of the measurement.  
The ground connection.  
A poorly located ground connection allows ground loops to add to the common mode  
signal.  
Frequency.  
Frequency is the most important factor in CMRR only because all of the factors  
mentioned above are frequency dependent. The unbalances of capacitance and  
inductance are more important as frequency increases. Therefore, good high-  
frequency practice is important when using a high impedance differential probe.  
On the other hand, if the differential probe is ac coupled to the circuit under test (the  
ac adapter is being used) the CMRR will be degraded below a certain frequency; the  
lower the frequency the worse the CMRR. This is because unbalance in the series  
capacitances of the ac coupler becomes more significant the lower the frequency.  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operating the Probe  
Recommended Test Equipment  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2
Calibration Tests and Adjustment  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Calibration Tests and Adjustment  
Equipment Required  
Introduction  
This chapter is divided into two sections. The first section gives  
calibration tests and the second adjustment procedures for the 1141A  
Equipment Required  
A complete list of equipment required for the calibration tests and adjustments is listed  
in “Recommended Test Equipment” on page 9. Equipment required for individual  
proceduresislistedatthe procedure. Anyequipmentsatisfying thecriticalspecifications  
listed may be substituted for the recommended model.  
The Test Board  
The test board is a supplied accessory for use during calibration tests and adjustments  
to connect signals to the differential probe (with or without adapters). A BNC connector  
connects the test board to a cable from the signal generator. The board includes a 50 Ω  
termination (two 100 resistors).  
C A U T I O N  
The power rating of the 50 termination is 1.0 W. Keep the signal input below 7 Vdc  
or rms to avoid degrading the termination.  
Once the probe tip caps and probe tips have been removed, the probe can be connected  
to the test board in one of three ways, as shown in the figure below. At each position of  
the probe, probe inputs are connected to a different combination of signal and ground.  
A separate terminal on the test board connects the ground of the probe to the signal  
ground.  
Figure 2-1  
Test Board Showing Probe Positions  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Calibration Tests and Adjustment  
The Test Board  
Calibration Tests  
These procedures test the probe’s electrical performance using applicable specifications  
given in “Performance Specifications and Characteristics” on page 55 as performance  
standards. Specifications applicable to individual tests are noted at the test for reference.  
Testing Interval  
The calibration testing procedures may be performed for incoming inspection of the  
instrument and should be performed periodically thereafter to ensure and maintain peak  
performance. The recommended test interval is yearly or every 2,000 hours of operation.  
Amount of use, environmental conditions, and the user’s experience concerning need  
for testing will contribute to verification requirements.  
Calibration Test Record  
The results of the calibration tests may be tabulated in the Calibration Test Record  
provided at the end of the calibration tests. The Calibration Test Record listed the  
calibration tests and provides an area to mark test results. The results recorded in the  
Calibration Test Record during initial inspection may be used for later comparisons of  
the tests during periodic maintenance, troubleshooting, and after repairs or adjustments.  
Calibration Test Procedures  
Procedures may be done individually or in any order.  
N O T E  
Allow the instrument to warm up for at least 30 minutes prior to beginning calibration  
tests.  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Calibration Tests and Adjustment  
dc Gain Accuracy  
dc Gain Accuracy  
This test checks the dc gain accuracy of the differential probe and the dc accuracy of  
the differential probe with attenuator and adapters.  
Specification: Probe alone, 2%; with attenuator adapter, 4%  
Equipment Required  
Recommended  
Equipment  
Required  
Agilent  
Critical Specifications  
Model/Part  
ac/dc Calibrator  
or  
dc Power Supply  
100 mV to 7 V  
E3632A  
100 mV to 7 V  
DVM  
Load  
0.5% accuracy, 10 µV resolution 3458A or  
E34401A  
BNC Feedthrough, 50 Ω  
Pasternack  
Enterprises  
PE6008-50 or  
Huber+Suhner  
22543742  
Cables (2)  
BNC 50 Ω  
10503A  
Adapters (2)  
Test Board  
BNC (f) to dual banana (m)  
No substitute  
1251-2277  
01141-66514  
Probe Gain Test Procedure  
1 Connect the probe and test equipment as shown in figure 2-2.  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Calibration Tests and Adjustment  
dc Gain Accuracy  
Figure 2-2  
2 Set up the 1142A probe control and power module as follows:  
b
Under DC Couple, press the Zero offset button.  
3 With the 1141A Probe Amp disconnected from the test PCA, adjust the Offset  
Null control on the 1142A until the DVM reads 0Vdc.  
If the probe output voltage cannot be set to 0V, subtract this voltage from the subsequent  
measurements in this test.  
4 Connect the input of the Probe Amp to the test board in the position shown in  
Figure 2-4 on page -31.  
5 Adjust the DC Source to output 100 mV (nom.)  
6 Record the V measurement from the top DVM in figure 2-2.  
in1  
7 Record the V  
measurement from the bottom DVM in figure 2-2.  
out1  
8 Connect the Probe Amp to measure a negative voltage as shown in figure 2-3.  
9 Record the V measurement from the bottom DVM in figure 2-2  
out2  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Tests and Adjustment  
dc Gain Accuracy  
Figure 2-3  
Signal to - input  
10 Calculate probe gain as  
(Vout Vout  
)
Vout  
-------------  
Vin  
1
2
= ---------------------------------  
2 × Vin  
1
Record the result of this calculation in the “Calibration Test Record” on page 39.  
To pass this test, the probe gain = 0.98 to 1.02  
N O T E  
N O T E  
Failure of the gain accuracy test can be caused by mis-adjustment of the probe. Perform  
the Probe Adjustment procedure in the Adjustments section later in this chapter and  
retest.  
10x Attenuator Accuracy Test  
If the gain test for the probe fails, the 10x Attenuator Accuracy Test will fail or the  
results will be poor. Do not continue until the probe passes the dc gain test.  
1 Disconnect the probe from the test board and connect the 10x attenuator  
adapter to the probe.  
2 Carefully connect the input of the probe/attenuator to the test board in the  
position shown in the figure below (signal to + input).  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Calibration Tests and Adjustment  
dc Gain Accuracy  
Figure 2-4  
Signal to + input  
3 Set the dc calibrator output 3 V dc.  
4 Record the V measurement from the top DVM in figure 2-2.  
in1  
5 Record the V  
measurement from the bottom DVM in figure 2-2  
out1  
6 Carefully connect the input of the probe/attenuator to the test board in the  
position shown in the figure below (signal to - input).  
Figure 2-5  
Signal to - input  
7 Record the V  
measurement from the bottom DVM in figure 2-2.  
out2  
8 Calculate the 10x attenuator gain as  
(Vout Vout  
)
Vout  
-------------  
Vin  
1
2
= ---------------------------------  
2 × Vin  
1
Record the result of this calculation in the “Calibration Test Record” on page 39.  
N O T E  
Failure of the accuracy test for the 10x attenuator can be caused by mis-adjustment of  
the low-frequency CMRR (LF CMRR) adjustment. Perform the Attenuator Adapter  
Adjustment procedure in the Adjustments section later in this chapter then retest the  
attenuator adapter. If if continues to fail, repair is necessary.  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Tests and Adjustment  
dc Gain Accuracy  
100x Attenuator Accuracy Test  
N O T E  
If the gain test for the probe fails, it will be reflected in the test for the 100x attenuator  
adapter. Do not continue until the probe passes the gain test.  
1 Disconnect the probe/attenuator from the test board. Remove the 10x  
attenuator adapter from the probe and connect the 100x attenuator adapter.  
2 Carefully connect the input of the probe/attenuator to the test board in the  
position shown in the figure below (signal to + input).  
Figure 2-6  
Signal to + input  
C A U T I O N  
Avoidexcessivepowerdissipationintheterminationonthetestboard. Keepthevoltage  
input at or below 7 Vdc.  
3 Set the dc source output to 7 Vdc.  
4 Record the V measurement from the top DVM in figure 2-2.  
in1  
5 Record the V  
measurement from the bottom DVM in figure 2-2  
out1  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Tests and Adjustment  
dc Gain Accuracy  
6 Carefully connect the input of the probe/attenuator to the test board in the  
position shown in the figure below (signal to - input).  
Figure 2-7  
Signal to - input  
7 Record the V  
measurement from the bottom DVM in figure 2-2.  
out2  
8 Calculate the 100x attenuator gain as  
(Vout Vout  
)
Vout  
-------------  
Vin  
1
2
= ---------------------------------  
2 × Vin  
1
Record the result of this calculation in the “Calibration Test Record” on page 39  
N O T E  
Failure of the accuracy test for the 100x attenuator can be caused by mis-adjustment  
of the low-frequency CMRR (LF CMRR) adjustment. Perform the Attenuator Adapter  
Adjustment procedure in the Adjustments section later in this chapter then retest the  
attenuator adapter. If if continues to fail, repair is necessary.  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Tests and Adjustment  
Bandwidth  
Bandwidth  
This test checks the high-frequency response of the 1141A Differential Probe. The  
bandwidth of the oscilloscope is characterized first so it is not a factor in the  
measurement.  
Specification (-3dB, dc coupled): dc to 200 MHz  
Equipment Required  
Equipment  
Required  
Critical Specifications  
Recommended  
Model/Part  
Oscilloscope  
Signal Generator  
Test Board  
Cable  
400 MHz bandwidth  
200 MHz at 230 mVrms  
No substitute  
54830A  
8648A  
01141-66504  
11500B  
Type N (m) 24-inch  
Type N (f) to BNC (m)  
Adapter  
1250-0077  
Procedure  
This test depends on the accuracy of the termination on the test board and the  
termination in the oscilloscope. Both should be with 1%.  
1 With the N cable and N-to-BNC adapter, connect the signal generator to the  
oscilloscope channel 1 input.  
2 Set the signal generator for 200 MHz at 0 dBm (about 224 mVrms).  
3 Set the 1142A front panel switches to Local and Zero offset.  
4 On the oscilloscope press AUTOSCALE, then set the following parameters.  
Menu  
Selection  
Setting  
TIMEBASE  
CHAN 1  
(time/div)  
2 ns/div  
(sensitivity)  
(input R)  
100 mV/div  
50 DC  
ACQUISITION  
Sampling Mode  
Memory Depth  
Sample Rate  
Averaging  
Real Time  
Automatic  
Automatic  
Enabled  
32  
# of avg  
5 The signal on screen should be about six divisions amplitude.  
Measure the peak-to-peak voltage of the channel 1 signal and record the  
reading.  
V
(1) = _____________ mV  
p-p  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Calibration Tests and Adjustment  
Bandwidth  
6 Reconfigure the equipment.  
a
b
c
Disconnect the signal generator cable from the oscilloscope input and connect it to  
the test board.  
Connect the output of the differential probe to the channel 1 input of the  
oscilloscope.  
Carefully connect the input of the probe to the test board in the position shown in  
the figure below (signal to + input).  
Figure 2-8  
Signal to + input  
7 Record the V reading on the oscilloscope.  
p-p  
V
(1) = _____________ mV  
p-p  
8 Divide the reading from step 6 by the reading from step 4.  
Answer from step 6  
Answer from step 4  
----------------------------------------------  
= ____________  
Record the result in the Calibration Test Record.  
The result should be 0.707 or greater, indicating a probe bandwidth of 200 MHz or more.  
The bandwidth can be checked at other signal levels. Change the signal generator output  
level and oscilloscope V/div range proportionally.  
N O T E  
Failure of the bandwidth test can be caused by mis-adjustment of the probe. Perform  
the Probe Adjustment procedure in the Adjustments section later in this chapter.  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Tests and Adjustment  
CMRR Test  
CMRR Test  
This test checks the CMRR at 1 MHz and 100 MHz.  
Specification 3000:1 at 1 MHz, 10:1 at 100 MHz  
Equipment Required  
Recommended  
Agilent  
Model/Part  
Equipment  
Required  
Critical Specifications  
Oscilloscope  
Signal Generator  
Test Board  
Cable  
400 MHz bandwidth at 1 mV/div 54830B  
1-100 MHz at = 400 mVrms  
No substitute  
8648A  
01141-66504  
11500B  
Type N (m) 24-inch  
Type N (f) to BNC (m)  
Adapter  
1250-0077  
Procedure  
1 Connect the probe power connector to the PROBE connection on the rear of  
the 1142A Probe Control and Power Module.  
2 Set the 1142A front panel switches to Local and Zero offset.  
3 Connect the probe output to the oscilloscope channel 1 input.  
4 Connect the input of the probe to the test board in the position shown in the  
figure below (signal to + input).  
Figure 2-9  
Signal to + input  
5 Connect the signal generator to the test board.  
6 Set the signal generator for 1 MHz at 385 mV  
(1 V ).  
p-p  
rms  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Calibration Tests and Adjustment  
CMRR Test  
7 On the oscilloscope, press AUTOSCALE and set the following parameters.  
Menu  
Selection  
Setting  
TIMEBASE  
CHAN 1  
(time/div)  
500 ns/div  
(sensitivity)  
(input R)  
200 mV/div  
50 DC  
ACQUISITION  
Sampling Mode  
Memory Depth  
Sample Rate  
Averaging  
Real Time  
Automatic  
Automatic  
Enabled  
32  
# of avg  
8 On the oscilloscope, measure the peak-to-peak voltage of the channel 1 signal,  
then V P-P, then press 1) and record the reading.  
V
(1) = _____________ mV  
p-p  
9 Connect the input of the probe to the test board in the position shown in the  
figure below (signal to both inputs).  
Figure 2-10  
Signal to both input  
10 Set the sensitivity to 1 mV/div.  
11 After the measurement settles (averaging is complete), record the V P-P  
reading.  
V
(2) = _____________ mV  
p-p  
12 Disconnect the probe amp from the test board and measure V  
on  
noise pp  
channel 1.  
13 Calculate the CMRR result as follows  
V
pp1  
CMRR = ------------------------------------  
V
Vnoisepp  
pp2  
14 The result in step 12 should be 3000, representing a CMRR of 3000:1 or more.  
Record the CMRR in the Calibration Test Record.  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Tests and Adjustment  
CMRR Test  
15 Connect the input of the probe to the test board in the position shown in the  
figure below (signal to + input).  
Figure 2-11  
Signal to + input  
16 Set the signal generator for 100 MHz at 0.0 dBm (about 224 mV , 632 mV ).  
rms  
p-p  
17 Set the oscilloscope to channel 1 and change the horizontal scale to 5 ns/div.  
18 After the measurement settles (averaging is complete), note the V P-P reading.  
V
(1) = _____________ mV  
p-p  
19 Connect the input of the probe to the test board in the position shown in the  
figure below (signal to both inputs).  
Figure 2-12  
Signal to both input  
20 Set the channel 1 sensitivity to 10 mV/div.  
21 After the measurement settles (averaging is complete), not the V P-P reading.  
V
(2) = _____________ mV  
p-p  
22 Disconnect the probe amp from the test board and measure V  
on  
noise pp  
channel 1.  
23 Calculate the CMRR result as follows  
V
pp1  
CMRR = ------------------------------------  
V
Vnoisepp  
pp2  
24 The result in step 21 should be 10, representing a CMRR of 10:1 or more.  
Record the CMRR in the Calibration Test Record.  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Tests and Adjustment  
Calibration Test Record  
Calibration Test Record  
1141A/1142A Differential Probe  
Tested by_________________________  
Work Order No.____________________  
Date____________________  
Serial No. ______________________________  
Recommended Test Interval - 1 Year/2000 hours  
Recommended next testing_________________  
Temperature_____________  
Test  
Limits  
Results  
dc Gain Accuracy  
Probe  
Only  
+0.98 mV to +1.02 mV  
_____________  
10x  
+0.096 mV to +0.104 mV  
+0.0096 mV to 0.0104 mV  
>0.707 at 200 MHz  
_____________  
_____________  
_____________  
_____________  
100x  
Bandwidth  
CMRR  
1 MHz  
3000:1  
10:1  
100 MHz  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Calibration Tests and Adjustment  
Probe Adjustment  
Adjustments  
This section provides adjustment procedures for the 1141A Differential Probe and  
attenuator adapters. There are no service adjustments for the 1142A Probe Control  
Module.  
Adjustment Interval  
None of the adjustment procedures that follow should be considered for a routine  
maintenance plan. The differential probe and attenuator adapters should be adjusted  
under conditions specified at the beginning of the respective procedures.  
N O T E  
Warm up the instrument for 30 minutes before starting adjustment procedures.  
Probe Adjustment  
This procedure adjust the high-frequency and low-frequency paths on the 1141A  
Differential Probe.  
N O T E  
Do not perform this procedure as a part of routine maintenance. Perform the procedure  
only if the probe does not meet specifications or has been repaired.  
C A U T I O N  
You are going to remove the covers of the probe, so the assembly inside will be exposed  
while under power. The PC assembly will be electrically and mechanically vulnerable.  
Do these adjustment procedures in an ESD-safe area.  
Avoid inadvertent contact between the powered assembly and nearby tools and  
equipment.  
Avoid mechanical damage by carefully handling the exposed assembly and cables.  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Calibration Tests and Adjustment  
Probe Adjustment  
The following equipment is required for this procedure. Procedures are based on the  
model or part number recommended.  
Equipment Required  
Equipment Required  
Critical Specifications  
Recommended  
Agilent  
Model/Part  
Function Generator  
Oscilloscope  
Signal Generator  
Test Board  
2.5 kHz, 1 V  
33120A  
p-p  
300 MHz at 1 mV/div  
200 MHz at 300 mVrms  
No substitute  
54830B  
8648A  
01141-66504  
10503A  
Cables (2)  
BNC, 50 Ω  
Adapter  
Type N (m) to BNC (f)  
1250-0780  
Probe Preparation  
The probe cover must be removed before adjustment. Drift due to temperature  
differences with and without covers is negligible.  
1 Remove the probe tip caps and probe tips.  
2 Loosen the probe clamp ring at the cable end of the probe (1/4 turn counter-  
clockwise) and slide it down the cable.  
3 Remove the bottom cover.  
a
At the cable end of the probe, separate the covers about centimeter (1/2 inch).  
b
Slide the bottom cover toward the cable end of the probe until the locator pins at  
the probe input clear the holes. Then, remove the cover.  
N O T E  
Note the position of the ground block at the input end of the probe. The ground block  
is held, through the PC assembly, by the grounding screw on the top of the probe. The  
ground block must be reinstalled on the PC assembly after the top cover is removed.  
Handle the PC assembly by the edges of the PC board.  
4 Remove the ground connection screw on the top of the probe.  
The ground block will become free.  
At the cable end, the PC board fits over pins inside the top cover.  
5 Lift the board off of the pins in the cover and slide it in the direction of the  
cable until the input connectors clear the front of the probe.  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Tests and Adjustment  
Adjustment Procedure  
6 As shown in the figure below, use the grounding screw to reinstall the ground  
block on the PC assembly.  
Figure 2-13  
Attaching Ground Block to Probe PC Assembly  
The ground block provides a mechanical and electrical connection when the probe PC  
assembly is connected to the test board.  
7 Connect the probe power connector to the PROBE connection on the rear of  
the 1142A Probe Control and Power Module.  
8 Connect the mains power to the 1142A.  
9 Set the 1142A front panel switches to Local and Zero offset.  
Adjustment Procedure  
Unless specified elsewhere, the procedures must be followed in the order given.  
The only adjustment which may be done separately is HF COMP, the high-frequency  
compensation.  
HF Gain and HF CMRR  
This adjustment sequence adjusts the HF Gain for unity gain at 500 kHz and the HF  
CMRR for minimum with a 500 kHz common mode signal.  
1 Set up the function generator.  
Sine wave  
500 kHz  
600 mV  
p-p  
2 Use BNC cables to connect the function generator to the oscilloscope.  
Generator OUTPUT to oscilloscope channel 1 input  
Generator TRIG OUTPUT to oscilloscope EXT TRIG  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Calibration Tests and Adjustment  
Adjustment Procedure  
3 On the oscilloscope, then press CLEAR DISPLAY. Press AUTOSCALE, then set  
up the following parameters.  
Menu  
Selection  
Setting  
TIMEBASE  
CHAN 1  
(time/div)  
500 ns/div  
(sensitivity)  
(input R)  
100 mV/div  
50 DC  
TRIG  
(mode)  
source  
level  
trg’d  
EXT  
1.00000 V  
ACQUISITION  
Sampling Mode  
Memory Depth  
Sample Rate  
Averaging  
Real Time  
Automatic  
Automatic  
Enabled  
32  
# of avg  
4 On the oscilloscope, measure the peak-to-peak voltage of the channel 1 signal  
and record the reading.  
V
(1) = _____________ mV  
p-p  
5 Disconnect the BNC cable from the channel 1 input and connect it to the BNC  
connector on the test board.  
6 Connect the output of the probe to the channel 1 input.  
7 Carefully connect the input of the probe to the test board in the position shown  
in the figure below (signal to + input).  
Figure 2-14  
Signal to + input  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Tests and Adjustment  
Adjustment Procedure  
8 Center adjustment R11, HF CMRR (see following figure).  
Figure 2-15  
R11, HF CMRR Adjustment  
9 Adjust R9, HF GAIN so the V (1) measurement is the same as in step 4, within  
p-p  
0.5%.  
Make the adjustment slowly so the oscilloscope display has time to react to signal  
averaging. Press CLEAR DISPLAY occasionally to restart averaging, which gives a  
quicker indication of changes.  
10 Carefully connect the probe to the test board in the positionshown in the figure  
below (signal to both inputs).  
Figure 2-16  
Signal to both inputs  
11 Set the function generator output to 1.0 V  
.
p-p  
12 On the oscilloscope, set the channel 1 sensitivity to 1.00 mV/div.  
13 Adjust R11 for minimum signal amplitude as shown in V reading for  
p-p  
channel 1. Adjust R11 slowly and use CLEAR DISPLAY frequently to restart  
averaging.  
Low Frequency Response and CMRR  
This adjustment sequence continues from the HF Gain and HF CMRR adjustments.  
Adjust R14 and C4 for pulse response, and adjust C6 for low-frequency CMRR.  
1 Change the function generator settings to:  
Square wave  
2.5 kHz  
600 mV  
p-p  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Tests and Adjustment  
Adjustment Procedure  
2 Change the oscilloscope settings to:  
Menu  
Selection  
(time/div)  
(sensitivity)  
Setting  
TIMEBASE  
CHAN 1  
50 µs/div  
100 mV/div  
3 Carefully connect the input of the probe to the test board in the position shown  
in the figure below (signal to + input).  
Figure 2-17  
Signal to + input  
4 Adjust R14 (LF Gain) and C4 (+ LF BANDWIDTH) for the flattest pulse top  
(see figure below). Again, adjust slowly and press CLEAR DISPLAY frequently  
to restart averaging.  
Figure 2-18  
R14 and C4 Adjustment  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Tests and Adjustment  
Adjustment Procedure  
5 Carefully connect the probe to the test board in position shown in the figure  
below (signal to both inputs).  
Figure 2-19  
Signal to both inputs  
6 Change the function generator settings to:  
Sine wave  
4 kHz  
1.0 V  
p-p  
7 Change the oscilloscope settings to:  
Menu  
Selection  
(time/div)  
(sensitivity)  
Setting  
TIMEBASE  
CHAN 1  
50 µs/div  
2 mV/div  
8 Adjust C6 (-LF BANDWIDTH) for minimum signal amplitude on the  
oscilloscope. Again, adjust C6 slowly and press CLEAR DISPLAY frequently to  
restart averaging.  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Tests and Adjustment  
Adjustment Procedure  
High Frequency Compensation  
This adjustment sequence continues from the Low Frequency Response and CMRR  
adjustment. However, it can be done separately if the probe meets all specifications  
except bandwidth. Adjust R13 for unity gain at 200 MHz.  
1 Connect the signal generator to the test board and set it for 200 MHz and  
300mV (107 mV ).  
p-p  
rms  
2 Carefully connect the input of the probe to the test board in the position shown  
in the figure below (signal to + input).  
Figure 2-20  
Signal to + input  
3 Press AUTOSCALE, then measure the peak-to-peak voltage on channel 1  
(Press SHIFT (blue), press V P-P, then press 1).  
4 Adjust R13 (HF COMP) to make the signal amplitude measurement on the  
oscilloscope 300 mV , or as close to that as possible. Adjust slowly and press  
p-p  
CLEAR DISPLAY frequently to restart averaging.  
Figure 2-21  
R13 Adjustment  
The minimum allowable amplitude is 212 mV . Typical values will be between 275 and  
p-p  
325 mV . the probe needs repair if the minimum cannot be reached.  
p-p  
Probe Reassembly  
1 Disconnect the probe power cable at the rear panel of the 1142A  
2 Remove the probe PC assembly from the test board. Be sure the probe input  
connectors remain attached to the probe.  
3 Remove the grounding screw and ground block from the PC assembly.  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Tests and Adjustment  
Adjustment Procedure  
4 Assemble the PC assembly in the top cover.  
The side of the PC assembly with the large hybrid is exposed when the assembly is in  
the top cover. The figure below shows how the top cover, PC board, and ground block  
fit together.  
Figure 2-22  
Reassembling the Probe  
a
Insert the input connectors first, and seat the cable end of the PC assembly over the  
pins at the rear of the cover.  
b
c
Position the ground block at the center-front of the PC assembly.  
Insert the grounding screw through the top cover and PC assembly and screw it into  
the ground block as shown.  
5 Replace the bottom cover.  
Position the cable strain relief and with one hand, hold the cable and top together.  
a
The flange on the strain relief has a notch that fits around a protrusion in the top cover.  
C A U T I O N  
Note where the two pins at the rear of the top cover enter the holes in the PC assembly.  
Position the cable wires away from these two areas. Otherwise, when the bottom cover  
is closed, part of it will pinch wires that are laying over these areas.  
b
c
Insert the pins at the front of the bottom cover into the holes at the front of the top  
cover.  
Close the two covers together and fasten with the probe clamp ring.  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Tests and Adjustment  
Attenuator Adapter Adjustment  
Attenuator Adapter Adjustment  
The following procedure should be used if it is necessary to adjust an attenuator adapter.  
Attenuator adapters have only characteristics; they do not have any specifications. An  
adapter will need adjustment only if one or more of the following occurs.  
If an adapter is to be used on a different 1141A probe that it was calibrated with  
last.  
If an adapter needs to be optimized to requirements for a special measurement.  
If an adapter is suspected of needing adjustment.  
N O T E  
Attenuator adapters should not be adjusted as part of routine maintenance. The design  
necessary to give the high CMRR and high impedance of the attenuators makes  
adjustment delicate.  
Additionally, because of broadband noise, a spectrum analyzer should be used to verify  
adjustment quality.  
Each attenuator contains two identical voltage dividers on a ceramic substrate. There  
are three adjustments. The low-frequency CMRR adjust the low-frequency balance  
between the two voltage dividers. The two high-frequency adjustments are electrically  
identical. Each adjusts the high-frequency compensation of one of the voltage dividers.  
To meet both pulse response and CMRR characteristics they are adjusted differently.  
Briefly, the adjustment procedure is:  
a
Adjustthepositive high-frequencyresponse(+HFRESP)forthe best pulse response  
using a 3.5 kHz square wave.  
b
Adjust the negative high-frequency response (-HF RESP) and Low-frequency CMRR  
(LF CMRR) for best CMRR using a 3.5 kHz square wave.  
The following equipment is required for this procedure. Procedures are based on the  
model or part number recommended.  
Equipment Required  
Equipment Required  
Critical Specifications  
3.5 kHz, 3 V to 16 V  
Recommended  
Model/Part  
Function Generator  
Oscilloscope  
Probe/Power Module  
Test Board  
33120A  
p-p  
100 MHz at 1 mV/div  
No substitute  
54830B  
1141A/1142A  
01141-66504  
10502A  
No substitute  
Cable  
BNC, 50 9-inch  
BNC, 50 36-inch  
Cable  
10503A  
Alignment tool  
Flat blade (supplied accessory) 8710-1961/  
Sprague-Goodman  
part, GTT-5G  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Calibration Tests and Adjustment  
Attenuator Adapter Adjustment  
Adjustment Procedure  
N O T E  
The attenuator must be adjusted when installed on the 1141A probe with which it will  
be used. The specifications and characteristics will not be met if the attenuator adapter  
is adjusted with one differential probe then used with another.  
1 Removetheprobepinsfromtheattenuatoradapteranddifferentialprobe, then  
attach the adapter to the probe.  
2 Set the 1142A front panel switches to Local and Zero offset.  
3 Use the 9-inch BNC cable to connect the function generator to the test board.  
The short cable minimized ground-loop voltages.  
4 Set up the function generator.  
Square wave  
3.5 kHz  
3.0 Vp-p for 10x adapter and 16 V for a 100x adapter.  
p-p  
5 Use the long BNC cable to connect the Trig Out of the function generator to  
the EXT TRIG of the oscilloscope.  
6 Set up the oscilloscope, then set the following parameters.  
Menu  
Selection  
Setting  
TIMEBASE  
CHAN 1  
(time/div)  
50 µs/div  
(sensitivity)  
(input R)  
50 mV/div  
50 DC  
TRIG  
(mode)  
source  
level  
trg’d  
EXT  
1.00000 V  
ACQUISITION  
Sampling Mode  
Memory Depth  
Sample Rate  
Averaging  
Real Time  
Automatic  
Automatic  
Enabled  
32  
# of avg  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Calibration Tests and Adjustment  
Attenuator Adapter Adjustment  
7 Connect the adapter/probe combination to the test boardinthe positionshown  
in the figure below.  
Figure 2-23  
Signal to + input  
8 Adjust the + HF RESP for best overall pulse response, the flattest pulse top.  
Use the figure below for adjust locations.  
Figure 2-24  
Adjustment Locations  
9 Change the function generator to 10 V (10x adapter adjustment only).  
p-p  
10 On the oscilloscope, press CHAN and set the sensitivity to 1 mV/div.  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Tests and Adjustment  
Attenuator Adapter Adjustment  
11 Connect the adapter/probe combination to the test boardinthe positionshown  
in the figure below (signal to both inputs).  
Figure 2-25  
Signal to both input  
12 Alternately adjust the LF CMRR and then the -HF RESP for a minimum signal  
on the oscilloscope. Repeat the adjustments until the signal is optimized to a  
minimum. Each adjustment should be set to minimize the component of the  
signal it affects most. Some high-frequency components of the signal are not  
affected by either adjustment.  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3
Service  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Service  
Introduction  
Introduction  
This section provides troubleshooting, service, and repair information for the  
1141A Differential Probe and 1142A Probe Control and Power Module. The  
troubleshooting information is provided to isolate a faulty assembly. When a  
faulty assembly has been located, the disassembly/assembly procedures help  
direct replacement of the assembly.  
W A R N I N G  
Maintenance should be performed by trained service personnel aware of the hazards  
involved (for example, fire and electric shock). When maintenance can be performed  
without power applied, the power cord must be removed from the instrument.  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Performance Specifications and Characteristics  
Performance Specifications and Characteristics  
The following table gives performance specifications used to test the 1141A and 1142A.  
It also gives performance characteristics that are typical for the probe system.  
Performance Specifications and Characteristics  
Parameter  
Probe alone  
With 10x attenuator  
With 100x attenuator  
SPECIFICATIONS  
1
Bandwidth (-3 dB, dc coupled) dc to 200 MHz  
Rise Time: (calculated)  
Gain Accuracy  
CMRR  
1.75 ns  
2.0%  
4.0%  
500 V(dc + peak ac)  
See the graphs in figure 3-1  
200 V(dc + peak ac)  
Maximum Input Voltage  
(see figure 3-2)  
Differential Input Range  
DC mode with no offset  
300 mV peak  
3.0 V peak  
30 V peak  
with DC Reject or appropriate 20 Vdc, decreasing to 200 Vdc, decreasing to 500 Vdc, decreasing to  
offset  
300 mV at 30 Hz  
3.0 mV at 30 Hz  
30 mV at 30 Hz  
Common-mode Operating  
Range  
dc  
20 Vdc  
linear change  
0.5 V  
200 Vdc  
linear change  
5 V  
500 Vdc  
linear change  
50 V  
dc to 30 Hz  
30 Hz to 200 MHz  
dc Offset Range  
20 V  
200 V  
500 V  
Input Impedance Resistance  
Capacitance  
1 MΩ  
7 pF  
9 MΩ  
3.5 pF  
10 MΩ  
2 pF  
ac Low-freq. Response (-3dB)  
dc Reject Response  
Output Impedance  
Thermal drift  
15 Hz  
1.5 Hz  
1.5 Hz  
5 Hz, 0.5 Hz, or 0.05 Hz (selectable irrespective of attenuator)  
50 Ω  
50 µVdc/°C  
Displayed noises  
50 µV  
rms  
Overload Recovery  
< 1 ms from overdrive that is less than the common mode range  
Note: 1. For maximum signal fidelity above 100 MHz, limit the probe input (without attenuators to 300 mV  
peak-to-peak.  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                                     
Service  
Performance Specifications and Characteristics  
Figure 3-1  
CMRR Specifications and Characteristics  
Legend  
A.  
B.  
C.  
D.  
E.  
CMRR specification for probe with no input adapters.  
Low-frequency CMRR specification for probe with the ac adapter.  
Typical CMRR characteristic for differential probe with no input adapters  
Typical CMRR characteristic for differential probe with 100x attenuator adapter at input.  
Typical CMRR characteristic for differential probe with 100x attenuator adapter at input.  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service  
Performance Specifications and Characteristics  
Figure 3-2  
Maximum Input Voltage vs. Frequency  
Legend  
A.  
B.  
C.  
Input voltage limits for probe alone.  
Input voltage limits for 10x adapter.  
Input voltage limits for 100x adapter.  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service  
General Characteristics  
General Characteristics  
Thefollowingcharacteristics apply tothe 1141A Differential Probewith the 1142AProbe  
Control and Power Module.  
Environmental Conditions  
Operating  
Non-operating  
Temperature  
Humidity  
0°C to +55 C° (32°F to +131°F)  
-40°C to +70°C (-40°F to +158°F)  
up to 95% relative humidity (non-  
condensing) at +40°C (+104°F)  
up to 90% relative humidity at +65°C  
(+149°F)  
Altitude  
up to 4,600 meters (15,000 ft)  
up to 15,300 meters (50,000 ft)  
Vibration  
Random vibration 5 to 500 Hz, 10  
minutes per axis, 0.3grms.  
Random vibration 5 to 500 Hz, 10 min.  
per axis, 2.41 grams. Resonant search  
5 to 500 Hz swept sine, 1Octave/min.  
sweep rate, (0.75g), 5 min. resonant  
dwell at 4 resonances per axis.  
Power  
Requirements  
Voltage: 90 to 132/198 to 264 Vac, 47 to 440 Hz  
Power: 25 VA maximum  
Weight  
Net: approximately 1.8 kg (4.0 lb.)  
Shipping: approximately 2.7 kg (6.0 lb.)  
Dimensions  
Refer to the outline drawings below.  
Figure 3-3  
1142A Probe Control  
and Power Module  
1141A Differential  
Probe  
Mechanical Dimensions  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Service  
Theory of Operation  
Theory of Operation  
The following discussion covers block-level theory for the 1141A/1142A differential  
probe system. Refer to the block diagram below.  
The differential probe system consists of two units, the 1141A Differential Probe with  
its accessories and the 1142A Probe Control and Power Module. For purposes of the  
following discussion, these will be called the probe and the control module respectively.  
Figure 3-4  
Differential Probe System Block Diagram  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Service  
Theory of Operation  
Differential Probe  
The probe contains a two-path differential amplifier with unity gain. It is implemented  
on a double-sided surface-mount PC board with the high-frequency path on one side and  
the low-frequency path on the other. The two paths are split directly after the differential  
input connections.  
High-Frequency Path  
The positive and negative inputs are ac-coupled at 33 Hz into identical impedance  
converters. The HF CMRR adjustment balances the gain at the outputs of the impedance  
converters. The impedance feed a semi-custom differential amplifier. An additional  
negative input to the differential amplifier brings in the sum of the low-frequency and  
feedbacksignals.TheFREQCOMPadjustmentprovidesvariablehigh-frequencypeaking  
of the differential amp. The output amp provides two signals. A feedback signal is  
summed with the low-frequency signals and the output signal is the final output of the  
probe. Overall probe gain is set by the HIGH FREQ GAIN adjustment.  
The entire signal portion of the high-frequency path is implemented on a hybrid IC.  
Support circuitry includes bias for the impedance converters and a bias supply for  
current sources on the hybrid.  
Low-Frequency Path  
The bandwidth of the low-frequency path is approximately 75 KHz. The positive and  
negative inputs are dc coupled into identical inverting op-amps with gain of 0.5. They  
provide a precision 1 M input impedance for the probe. The LOW FREQ BANDWIDTH  
adjustments match the gain and phase of the low-frequency path to that of the feedback  
from the probe output. One of the two adjustments is set to match the properties of the  
feedback and the other is to match the two low-frequency paths. These adjustments  
affect the CMRR quality of the probe. The inputs of the inverting op-amps include  
protection for ESD and over-voltage conditions.  
The inverting op-amps feed a precision differential amplifier with unity gain and a single-  
ended output. The output is fed to the summing amp and to the control module for use  
in the dc reject circuit.  
Summing Amp  
The summing amp combines the feedbacksignal, the low-frequencysignal, andthe offset  
signal. The LOW FREQ GAIN adjustment matches the gain of the low-frequency path to  
the overall gain.  
Control and Power Module  
The control and power module provides offset functions, local and remote control, and  
power to the probe system.  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Service  
Theory of Operation  
Offset Functions  
There are two offset functions developed in the control module: variable offset and dc  
reject. A variable offset voltage with coarse and fine adjustments can be selected by the  
front panel controls. The offset level is buffered by U8 and selected by multiplexer U3  
as the input to offset amp U7. The output of the offset amp is summed with the low-  
frequency signal and feedback which gives dc coupling in the probe. Front panel  
screwdriver adjustment Offset Null zeros the dc output from the probe when the dc input  
and offset are zero.  
For dc Reject, an output from the low-frequency amplifier in the probe (LFSIG) is used  
to develop a voltage used to null the dc component of the input signal. LFSIG is an input  
to U6, an inverting amplifier and low-pass filter. Multiplexer U3 selects one of three  
capacitors to set a roll-off frequency of 0.05, 0.5, or 5 Hz. The output of U6 is selected,  
again by U3, as the input to the offset amp. When the output of the offset amp is summed  
into the low frequency path, the result is cancellation of the dc component of the input  
signal. Front panel screwdriver adjustment DC Reject Gain adjusts the gain of the dc  
reject circuit.  
Local and Remote Control  
The front panel switch controls the dc reject and offset functions. It also selects remote  
operation, which allows control through the rear panel remote input connector.  
Power Supply  
The supply provides 6 V and 15 V for the probe and analog control circuitry as well  
as +5 V for the digital control circuitry.  
Attenuator Adapters  
The 10x and 100x Attenuator Adapter are similar. A ceramic substrate carries two  
attenuators, one for each input polarity. A variable resistor adjusts the low-frequency  
balance (LF CMRR) between the two attenuators. The high-frequency adjustments are  
the same for each attenuator. Each attenuator is adjusted differently. One attenuator is  
adjusted for optimum pulse response and the other for best high-frequency CMRR.  
Figure 3-5  
Attenuator Adapter, Simplified Schematic  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Service  
Theory of Operation  
Test Board  
The test board is a device for conveniently connecting test signals to the differential  
probe. The probe can be connected to the board with the signal to the positive, negative,  
or both inputs.  
Figure 3-6  
Test Board Schematic  
62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Service  
Service Policy  
Service Policy  
For parts of the 1141A/1142A probe system that are complex, the service policy is for  
assembly-level repair. For parts of the system with simple circuitry, the service policy is  
component-level repair.  
The service policy for the 1141A Differential Probe is assembly-level repair. Assemblies  
include the PC assembly and cable. The PC assembly is an “exchange assembly.” A  
repaired and tested assembly is shipped upon receipt of the defective assembly.  
The attenuator and ac coupling adapters are shipped as complete assemblies.  
The service policy for the 1142A Probe Control and Power Module is component-level  
repair.  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Service  
Troubleshooting  
Troubleshooting  
Usethefollowingparagraphstoassistintroubleshootingproblemswiththe1141A/1142A  
Differential Probe.  
Probe Troubleshooting  
To troubleshoot the probe:  
1 Apply a known signal to the input of the probe.  
2 Check for an identical output at the output coax to the cable. This connection  
is the one soldered to the PC board. If the probe output cable is not terminated,  
or the coax is open, the output signal will be about twice the amplitude of the  
input signal.  
3 Ifthesignalisincorrect, checkthepowersupplyvoltagesfromthe1142AProbe  
Control and Power Module. Use the cable diagram on the next page.  
4 Troubleshoot the cable with an ohmmeter. Use the cable diagram on the next  
page.  
Probe Control and Power Module  
Troubleshooting  
The circuitry consists of simple power supplies, operational amplifiers, and TTL. Use  
conventional troubleshooting techniques. A complete parts list, component locator, and  
schematics are provided later in this chapter.  
64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Service  
Troubleshooting  
Figure 3-7  
1142A Probe Control and Power Module  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Removing and Replacing Assemblies  
Removing and Replacing Assemblies  
This section contains procedures for the removal and replacement of major assemblies.  
C A U T I O N  
Never remove or install any assembly with the instrument power ON. Component  
damage can occur.  
Differential Probe  
Use the following procedure to remove and replace the amplifier PC board in the  
differential probe.  
C A U T I O N  
C A U T I O N  
ELECTROSTATIC DISCHARGE can damage electronic components. Use grounded  
wrist straps and mats when servicing the probe.  
Handle the differential probe carefully once it has been disassembled. If unsupported,  
the weight of the cable can put strain on the PC board.  
Disassemble Probe  
1 Remove the probe tip caps and probe tips.  
2 Loosen the probe clamp ring at the cable end of the probe (1/4 turn counter-  
clockwise) and slide it down the cable.  
3 Remove the bottom cover.  
a
At the cable end of the probe, separate the covers about one centimeter (1/2 inch).  
b
Slide the bottom cover toward the cable end of the probe until the locator pins at  
the probe input clear the holes. Then, remove the cover.  
Note the way the cable strain relief is keyed and held at the rear of the top cover.  
4 Remove the probe top cover.  
The ground screw passes through the top cover and PC board and screws into the ground  
block.  
a
Remove the ground screw on the top of the probe.  
At the cable end, the PC board fits over the pins in the top cover.  
b
Lift the board off of the pins and slide it in the direction of the cable until the input  
connectors clear the front of the probe.  
5 Un-solder the two connections where the coaxial output cable connects to the  
PC board.  
6 Disconnect the cable connector from the probe PC board.  
66  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Service  
Removing and Replacing Assemblies  
Reassemble Probe  
The ground screw passes through the top cover and PC board and screws into the ground  
block.  
1 If replacing the PC board, remove the input connectors from the old board and  
put them on the new one.  
2 If replacing the cable, note the orientation of the probe clamp ring on the old  
cable, remove the ring and put it on the new cable.  
3 Connect the cable connector to the PC board.  
4 Solder the two connections of the coaxial cable to the PC board.  
On a new probe cable, the conductors of the coaxial cable are connected by a heavy  
single wire. Cut the heavy wire so it matches the wire on the cable that was removed.  
5 Assemble the PC assembly into the top cover.  
The large hybrid is exposed when the assembly is in the top cover. The figure on the  
below shows the sequence of the ground screw, top cover, PC board, and ground block.  
Figure 3-8  
Reassembling the Probe  
a
Insert the input connectors first, and seat the cable end of the PC assembly over the  
pins at the rear of the cover.  
b
c
Position the ground block at the front of the PC assembly.  
Insert the grounding screw through the top cover and screw it into the ground block  
as shown in the figure above.  
6 Replace the bottom cover.  
a
Position the cable strain relief and with one hand, hold the cable and top cover  
together.  
The flange on the strain relief has a notch that fits over a protrusion in the top cover.  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Removing and Replacing Assemblies  
C A U T I O N  
Note where the two pins at the rear of the top cover enter the holes in the PC assembly.  
Position the cable wires away from these two areas. When the bottom cover is closed,  
part of it will pinch wires that are laying over these areas.  
b
c
Insert the pins at the front of the bottom cover into the holes at the front of the top  
cover.  
Close the covers together and fasten with the probe clamp ring. If the covers do not  
fit together tightly and easily, check for pinched wires (see caution above).  
Probe Adapters  
Use the following procedure to disassemble the probe adapters. The adapter housing  
consists of two plastic parts, one of which slides into the other. The parts are held  
together by the spring effect of two plastic tabs on the inner part.  
Mechanically, the ac adapter is about the same as the 10x and 100x attenuator adapters.  
The attenuator adapters have an extra ground connector which connects the substrate  
ground to the thumbwheel screw and plating inside the housing.  
Disassemble Adapter  
1 Remove the probe tip caps and probe tips from the adapter input.  
2 Hold the adapter in one hand taking care not to block the output end of the  
adapter (the end which attaches to the probe).  
3 Note the view of the input end of the adapter in the figure below. The arrows  
indicate the holding tabs.  
Figure 3-9  
Disassembling Adapters  
4 With the thumb and forefinger, squeeze the tabs together, as indicated by the  
arrows. Simultaneously, push the tabs into the outer housing so the inner  
housing begins to slide out.  
5 While holding the outer housing, push back against the thumbwheel until the  
inner housing can be grasped and removed.  
Reassemble Adapter  
Reassembling the adapter is slightly harder because you have to align the connector pins  
and thumbwheel screw, while sliding the inner housing and outer housing together.  
1 Be sure the input connectors and output pins are present and seated on the  
substrate or PC board.  
68  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Service  
Removing and Replacing Assemblies  
2 Combine the thumbwheel screw and thumbwheel and insert them into the hole  
in the outer housing.  
3 Insert the substrate/board into the outer housing. Slip the attenuator ground  
(attenuator adapters only) over the thumbwheel screw and seat the input  
connectors in the proper holes in the housing.  
Figure 3-10  
Reassembling the Adapter  
Do not force reassembly of the adapter. The housing halves will slide together with  
moderate friction.  
4 Align the inner housing tabs with the grooves in the side of the outer housing  
and slide the two partly together.  
While seating the thumbwheel screw and pins, it will help to hold the assembly vertically,  
with the input end of the adapter down. This will allow the assemblies to sit vertically  
and more easily align with the holes in the inner housing.  
5 Seat the thumbwheel screw first, then the output pins, into the appropriate  
holes in the inner housing.  
6 Once the two housings are nearly together, press them together firmly until  
the tabs click into place.  
69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service  
Removing and Replacing Assemblies  
Probe Control and Power Module  
Use the following procedure to disassemble the probe control and power module.  
W A R N I N G  
W A R N I N G  
Hazardous voltages exist on the power supply. To avoid electrical shock, adhere closely  
to the following procedures.  
1 Remove the power cord.  
2 Remove four flathead screws and remove the top cover.  
Be sure to reconnect the safety ground when reassembling the instrument.  
3 Unplug the safety ground from the tab on the rear panel of the instrument.  
4 Note the orientation of the knobs. Remove the two knobs.  
5 On the bottom of the instrument, remove the 5mm screw that fastens the  
transformer support.  
6 Remove the following pan-head screws.  
Three directly on the PC board.  
Two on the ac input connector.  
Two on the heatsink.  
7 Remove the PC board. Slide it slightly forward so parts will clear the rear panel,  
then lift the rear of the board out while sliding it backwards.  
8 Remove the two heatsink spacers from the standoffs that were directly under  
the heatsink.  
9 Reverse the procedure to reassemble the control and power supply.  
70  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Service  
Replaceable Parts  
Replaceable Parts  
This section contains information for ordering parts. Service support for the 1141A  
DifferentialProbeistotheassemblylevel. Servicesupportfor the adaptersisas complete  
assemblies, except for the probe tip caps and the probe tips. Service support for the  
1142A Probe Control and Power Module is to the component level.  
Parts List  
The replaceable parts lists include all parts relevant to the applicable service levels. The  
information given for each part consists of the following:  
Reference designator  
Part number  
Total quantity (Qty) in instrument or on assembly. The total quantity is given once  
and at the first appearance of the part number in the list.  
Description of part  
Typical manufacturer of part in a five-digit code.  
Ordering Information  
To order a part in the material part list, quote the part number, indicate the quantity  
desired, and address the order to the nearest Agilent Technologies Sales Office.  
To order a part not listed in the material list, include the instrument part number,  
instrument serial number, a description of the part (including its function), and the  
number of parts required. Address the order to the nearest Agilent Technologies Sales  
Office.  
Direct Mail Order System  
Within the USA, Agilent Technologies can supply parts through a direct mail order  
system. There are several advantages to this system:  
Direct ordering and shipment from the Agilent Technologies parts center in  
California, USA.  
No maximum or minimum on any mail order (there is a minimum amount for parts  
ordered through a local Agilent Technologies Sale Office when the orders require  
billing and invoicing).  
Prepaid transportation (there is a small handling charge for each order).  
No invoices.  
In order for Agilent Technologies to provide these advantages, a check or money order  
must accompany each order.  
71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
               
Service  
Replaceable Parts  
Mail order forms and specific ordering information are available through your local  
Agilent Technologies Sales Office. Addresses and telephone numbers are located in a  
separate document shipped with the manuals.  
Manufacturers’ Codes  
A list of manufacturers’ codes is given the table below. The codes are given for parts in  
the parts lists. The table gives the manufacturer and address for each code.  
Manufacturers’ Code List  
Mfr. No.  
Name  
Address  
00000  
06665  
24546  
27014  
28480  
32997  
Any satisfactory supplier  
Precision Monolithics Inc.  
Corning Glass Works (Bradford)  
National Semiconductor Corp  
Agilent Technologies Corporate Hq  
Bourns Inc.  
Santa Clara, CA 95050  
Bradford, PA 16701  
Palo Alto, CA 94304  
Palo Alto, CA 94304  
Riverside, CA 92507  
72  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Service  
Replaceable Parts  
Exploded View  
Figure 3-11  
1141A Differential Probe Parts  
73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Replaceable Parts  
Figure 3-12  
Power and Control Board Component Locator  
74  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Replaceable Parts  
Figure 3-13  
Cabling Diagram  
75  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Replaceable Parts  
Figure 3-14  
IC Connectors not Shown  
Supply  
Pin No.  
IC Group  
+15  
-15  
GND  
+15  
-15  
NC  
+5  
GND  
14  
U3  
3
15  
7
U6-8, 11  
U9  
4
1, 5, 8  
16  
8
1142A Probe Control and Power Module  
76  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Replaceable Parts  
Figure 3-15  
1142A Probe Control and Power Module  
77  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Replaceable Parts  
1141A and 1142A Replaceable Parts  
Ref.  
Des.  
Part  
Number  
Qty  
Description  
Mfr.  
Code  
Mfr. Part  
Number  
1141A Differential Probe  
A1  
01141-66503  
01141-26102  
01141-22301  
01141-22401  
01141-24102  
01141-44702  
01141-44102  
01141-22502  
01141-94307  
1
2
1
1
2
1
1
1
1
PC ASSEMBLY-DIFFERENTIAL PROBE  
PROBING PIN  
28480 01141-66501  
28480 01141-26102  
28480 01141-22301  
28480 01141-22401  
28480 01141-24101  
28480 01141-44701  
28480 01141-44101  
28480 01141-22501  
28480 01141-94301  
E1  
E2  
GROUND BLACK  
E3  
GROUND SCREW  
MP1  
MP2  
MP3  
MP4  
MP5  
CAP-PROBING PIN  
PROBE BOTTOM COVER  
PROBE TOP COVER  
PROBE POS CLAMP RING  
DIFFERENTIAL PROBE LABEL  
W1  
01141-61603  
1
PROBE CABLE ASSEMBLY  
28480 01141-61601  
10X Attenuator Adapter  
PROBING PIN  
01141-26102  
01141-24102  
2
2
28480 01141-26102  
28480 01141-24101  
CAP-PROBING PIN  
100X Attenuator Adapter  
PROBING PIN  
01141-26102  
01141-24102  
2
2
28480 01141-26102  
28480 01141-24101  
CAP-PROBING PIN  
AC Adapter  
01141-26102  
01141-24102  
2
2
PROBING PIN  
CAP-PROBING PIN  
28480 01141-26102  
28480 01141-24101  
1142A PROBE CONTROL AND POWER MODULE  
78  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Service  
Replaceable Parts  
1141A and 1142A Replaceable Parts  
Ref.  
Des.  
Part  
Number  
Qty  
Description  
Mfr.  
Code  
Mfr. Part  
Number  
A1  
01142-66501  
0515-0374  
0515-1031  
0515-1579  
1
7
4
1
PC ASSEMBLY-POWER AND CONTROL  
SCREW-MACHINE M3 10mm-LG  
28480 01142-66501  
H1  
H2  
H3  
00000 ORDER BY DESP.  
00000 ORDER BY DESP.  
00000 ORDER BY DESP.  
SCREW-MACHINE M3 6mm-LG 90-DEG-FLH-HD  
SCREW-MACHINE M5 18mm-LG  
MP1  
MP2  
MP3  
MP4  
MP5  
MP6  
01142-47702  
01142-44101  
01142-24701  
0370-1097  
1
1
2
2
4
5
BOTTOM CHASSIS  
TOP CHASSIS  
SPACER-HEAT SINK  
KNOB-POINTER  
FOOT  
28480 01142-47701  
28480 01142-44101  
28480 01142-24701  
28480 0340-1097  
28480 0403-0727  
28480 5041-0234  
0403-1012  
5041-0234  
KEYCAP  
W1  
8120-1521  
1
POWER CORD 18-AWG 3-COND 90-IN-LG (US/  
Canada only)  
28480 8120-1521  
79  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service  
Replaceable Parts  
Power Supply Replaceable Parts  
Ref. Des.  
Part  
Number  
Qty  
Description  
Mfr.  
Code  
Mfr. Part  
Number  
Prefix the reference designators with A1  
CAPACITOR-FXD 0.33UF 10% 50VDC  
CAPACITOR-FXD 0.1UF 5% 100VDC MET-POLY  
CAPACITOR-FXD 2200UF+30-10% 50VDC AL  
CAPACITOR-FXD 2200UF+30-10% 50VDC AL  
CAPACITOR-FXD 0.01UF 10% 100VDC CER  
CAPACITOR-FXD 0.01UF 10% 100VDC CER  
CAPACITOR-FXD 0.01UF 10% 100VDC CER  
CAPACITOR-FXD 0.01UF 10% 100VDC CER  
CAPACITOR-FXD 4.7UF 10% 35VDC TA  
CAPACITOR-FXD 4.7UF 10% 35VDC TA  
NOT ASSIGNED  
C1  
0160-6190  
0160-5474  
0180-3298  
0180-3298  
0160-6500  
0160-6500  
0160-6500  
0160-6500  
0180-3845  
0180-3845  
2
1
2
28480 0160-6190  
28480 0160-5474  
28480 0180-3298  
28480 0180-3298  
28480 0160-6500  
28480 0160-6500  
28480 0160-6500  
28480 0160-6500  
28480 0180-3845  
28480 0180-3845  
C2  
C3  
C4  
C5  
4
C6  
C7  
C8  
C9  
4
1
C10  
C11  
C12  
C13  
C14  
C15  
C16  
C17  
C18  
C19  
C20  
C21-22  
C23  
C24  
C25  
C26  
0160-5581  
0160-6190  
0160-7060  
0160-4801  
0180-3845  
0180-3845  
CAPACITOR-FXD 0.033UF 10% 63VDC  
CAPACITOR-FXD 0.33UF 10% 50VDC  
CAPACITOR-FXD 3.3UF 10% 63VDC  
CAPACITOR-FXD 100PF 5% 100VDC CER  
CAPACITOR-FXD 4.7UF 10% 35VDC TA  
CAPACITOR-FXD 3.3UF 10% 63VDC  
NOT ASSIGNED  
28480 0160-5581  
28480 0160-6190  
28480 0160-7060  
28480 0160-4801  
28480 0180-3845  
28480 0180-3845  
1
1
0160-5471  
0160-5469  
1
1
CAPACITOR-FXD 0.1UF 5% 50VDC MET-POLY  
CAPACITOR-FXD 1UF 10% 50VDC MET-POLY  
NOT ASSIGNED  
28480 0160-5471  
28480 0160-5469  
0180-3784  
0180-3784  
0180-3784  
0180-3784  
4
CAPACITOR-FXD 22UF 20% 25VDC TA  
CAPACITOR-FXD 22UF 20% 25VDC TA  
CAPACITOR-FXD 22UF 20% 25VDC TA  
CAPACITOR-FXD 22UF 20% 25VDC TA  
28480 0180-3784  
28480 0180-3784  
28480 0180-3784  
28480 0180-3784  
80  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service  
Replaceable Parts  
Power Supply Replaceable Parts  
Ref. Des.  
Part  
Number  
Qty  
Description  
Mfr.  
Code  
Mfr. Part  
Number  
CR1  
NOT ASSIGNED  
CR2  
1901-1087  
1901-1087  
2
DIODE-PWR RECT 100V 3A  
DIODE-PWR RECT 100V 3A  
NOT ASSIGNED  
04713 MR501  
04713 MR501  
CR3  
CR4  
CR5  
1901-0731  
1901-0731  
8
8
DIODE-PWR RECT 400V 1A  
DIODE-PWR RECT 400V 1A  
NOT ASSIGNED  
28480 1901-0731  
28480 1901-0731  
CR6  
CR7-8  
CR9  
1901-0731  
1901-0731  
1901-0731  
1901-0731  
1901-0734  
1901-0734  
8
8
8
8
2
2
DIODE-PWR RECT 400V 1A  
DIODE-PWR RECT 400V 1A  
DIODE-PWR RECT 400V 1A  
DIODE-PWR RECT 400V 1A  
DIODE-PWR RECT 1N5818 30V 1A  
DIODE-PWR RECT 1N5818 30V 1A  
NOT ASSIGNED  
28480 1901-0731  
28480 1901-0731  
28480 1901-0731  
28480 1901-0731  
04713 IN5818  
CR10  
CR11  
CR12  
CR13  
CR14  
CR15-16  
CR17  
CR18  
04713 IN5818  
1901-0731  
1901-0731  
DIODE-PWR RECT 400V 1A  
DIODE-PWR RECT 400V 1A  
28480 1901-0731  
28480 1901-0731  
DS1  
1990-0521  
1
LED-LAMP LUM-INT=2.2MCD IF=50MA-MAX  
24840 5082-4955  
E1  
E2  
2110-0642  
2110-0565  
1
1
FUSEHOLDER 6.3A 250A  
FUSEHOLDER CAP  
28480 2110-0642  
28480 2110-0565  
F1  
2110-0201  
0515-1579  
01142-21101  
1
1
1
FUSE 0.25A 250V TD  
SCREW-MACHINE M5 18mm-LG  
HEAT SINK  
28480 2110-0201  
28480 0515-1579  
28480 01142-21101  
H1  
HS1  
81  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service  
Replaceable Parts  
Power Supply Replaceable Parts  
Ref. Des.  
Part  
Number  
Qty  
Description  
Mfr.  
Code  
Mfr. Part  
Number  
J1  
J2  
J3  
1252-4731  
1252-1487  
1252-3935  
1
1
1
CONNECTOR-AC PWR  
28480 1251-4743  
28480 1252-1487  
28480 1252-3134  
CONN-RECT D-SUBMIN 9-CKT (remote)  
CONNECTOR-ROUND 8-CKT (probe)  
MP1  
MP2  
MP3  
MP4  
MP5-6  
MP7  
MP8  
MP9  
1400-1604  
1205-0732  
0361-0685  
0340-1211  
1
4
3
3
LED MOUNT  
28480 1400-1604  
28480 1205-0732  
28480 0361-0685  
28480 0340-1211  
SPRING CLIP  
RIVET-BLIND DR-PIN RNDH 0.125DIA  
INSULATOR-THERMAL  
NOT ASSIGNED  
1400-0249  
1
1
1
CABLE TIE 0.062-0.625-DIA 0.091-WD NYL  
TRANSFORMER SUPPORT  
WASHER-TRANSFORMER SUPPORT  
16956 08-465/GRY  
28480 01142-24702  
28480 01142-28801  
01142-24702  
01142-28801  
R1-2  
R3  
NOT ASSIGNED  
0757-0442  
0757-0465  
0757-0199  
0757-0199  
0698-4431  
0699-1203  
0757-0434  
0698-4431  
0699-1203  
0757-0434  
0757-0420  
0698-4002  
0699-1203  
4
1
5
RESISTOR 10K 1% 0.125W TF TC=0 100  
RESISTOR 100K 1% 0.125W TF TC=0 100  
RESISTOR 21.5K 1% 0.125W TC=0 100  
RESISTOR 21.5K 1% 0.125W TC=0 100  
RESISTOR 2.05K 1% 0.125W TF TC=0 100  
RESISTOR 120.0 1% 0.125W TF TC=0 25  
RESISTOR 3.65K 1% 0.125W TF TC= 100  
RESISTOR 2.05K 1% 0.125W TF TC=0 100  
RESISTOR 120.0 1% 0.125W TF TC=0 25  
RESISTOR 3.65K 1% 0.125W TF TC=0 100  
RESISTOR 750 1% 0.125W TF TC=0 100  
RESISTOR 5K 1% 0.125W TF TC=0 100  
RESISTOR 120.0 1% 0.125W TF TC=0 25  
24546 CT4-1/8-TO-1002-F  
24546 CT4-1/8-TO-1003-F  
24546 CT4-1/8-TO-2152-F  
24546 CT4-1/8-TO-2152-F  
24546 CT4-1/8-TO-2052-F  
28480 0699-1203  
R4  
R5  
R6  
R7  
2
4
2
R8  
R9  
24546 CT4-1/8-TO-3651-F  
24546 CT4-1/8-TO-2052-F  
28480 0699-1203  
R10  
R11  
R12  
R13  
R14  
R15  
24546 CT4-1/8-TO-3651-F  
24546 CT4-1/8-TO-751-F  
24546 CT4-1/8-TO-5001-F  
28480 0699-1203  
1
2
82  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service  
Replaceable Parts  
Power Supply Replaceable Parts  
Ref. Des.  
Part  
Number  
Qty  
Description  
Mfr.  
Code  
Mfr. Part  
Number  
R16  
R17  
R18  
R19  
R20  
R21  
R22  
R23  
R24  
R25  
R26  
R27  
R28  
R29  
R30  
R30  
R32  
R33-39  
R40  
R41  
R42  
R43  
R44  
R45  
R46  
R47  
R48  
0698-6317  
0698-4002  
0699-1203  
0698-6317  
2
RESISTOR 500 0.1% 0.125W TF TC=0 25  
RESISTOR 5K 1% 0.125W TF TC=0 100  
RESISTOR 120.0 1% 0.125W TF TC=0 25  
RESISTOR 500 0.1% 0.125W TF TC=0 25  
NOT ASSIGNED  
28480 0698-6317  
24546 CT4-1/8-TO-5001-F  
28480 0699-1203  
28480 0698-6317  
0698-8827  
0757-0442  
0757-0427  
0757-0401  
2100-3161  
2100-3056  
0757-0199  
0757-0199  
0757-0442  
0757-0442  
0757-0280  
0757-0460  
1
RESISTOR 1M 1% 0.125W TF TC=0 100  
RESISTOR 10K 1% 0.125W TF TC=0 100  
RESISTOR 1.5K 1% 0.125W TF TC=0 100  
RESISTOR 100 1% 0.125W TF TC=0 100  
RESISTOR-TRMR 20K 10% TKF SIDE-ADJ  
RESISTOR-TRMR 5K 10% TKF SIDE-ADJ 17-TRN  
RESISTOR 21.5K 1% 0.125W TF TC=0 100  
RESISTOR 21.5K 1% 0.125W TF TC=0 100  
RESISTOR 10K 1% 0.125W TF TC=0 100  
RESISTOR 10K 1% 0.125W TF TC=0 100  
RESISTOR 1K 1% 0.125W TF TC=0 100  
RESISTOR 61.9K 1% 0.125W TF TC=0 100  
NOT ASSIGNED  
28480 0698-8827  
24546 CT4-1/8-TO-1002-F  
24546 CT4-1/8-TO-1501-F  
24546 CT4-1/8-TO-101-F  
32997 3006P-1-203  
1
1
1
1
32997 3006P-1-502  
24546 CT4-1/8-TO-2152-F  
24546 CT4-1/8-TO-2152-F  
24546 CT4-1/8-TO-1002-F  
24546 CT4-1/8-TO-1002-F  
24546 CT4-1/8-TO-1001-F  
24546 CT4-1/8-TO-6192-F  
1
1
0757-0123  
2100-4250  
2100-4250  
0757-0458  
0683-2755  
0698-4517  
0698-3271  
0698-8961  
0757-0199  
1
2
RESISTOR 34.8K 1% 0.125W TF TC=0 100  
RESISTOR-VAR 10K 20%  
28480 0757-0123  
28480 2100-4250  
RESISTOR-VAR 10K 20%  
28480 2100-4250  
1
1
1
1
1
RESISTOR 51.1K 1% 0.125W TF TC=0 100  
RESISTOR 2.7M 1% 0.25W TF TC=0 100  
RESISTOR 127K 1% 0.125W TF TC=0 100  
RESISTOR 115K 1% 0.125W TF TC=0 100  
RESISTOR 909K 1% 0.125W TF TC=0 100  
RESISTOR 21.5K 1% 0.125W TF TC=0 100  
24546 CT4-1/8-TO-5112-F  
28480 0683-2755  
24546 CT4-1/8-TO-1273-F  
24546 CT4-1/8-TO-1153-F  
28480 0698-8961  
24546 CT4-1/8-TO-2152-F  
83  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service  
Replaceable Parts  
Power Supply Replaceable Parts  
Ref. Des.  
Part  
Number  
Qty  
Description  
Mfr.  
Code  
Mfr. Part  
Number  
RP1  
1810-1242  
1
RESISTOR NETWORK  
28480 1810-1242  
S1  
S2  
3101-2609  
3101-3007  
1
1
SWITCH-SL DPST STD 5A 250VAC PC  
SWITCH-6 STATION ASSEMBLY  
28480 3101-2609  
28480 3101-3007  
T1  
9100-4750  
1
TRANSFORMER-POWER (with mtg. hardware)  
28480 9100-4750  
U1  
U2  
U3  
U4  
U5  
U6  
U7  
U8  
U9  
U10  
U1  
1826-1403  
1826-1670  
5081-9233  
1826-1403  
1826-1670  
1826-1381  
1826-1381  
1826-1381  
1820-3177  
1826-0774  
1826-0635  
2
2
1
IC V RGLTER-ADJ-POS 3/40V  
00000 LT317AT  
IC V RGLTR-ADJ-NEG -37/1.2V TO-220 PKG  
ANALOG MULTIPLEXER 4 CHNL 16 -DIP-P  
IC V RGLTR ADJ-POS 3/40V  
00000 LT337AT  
28480 1820-2182  
00000 LT317AT  
IC V RGLTR-ADJ-NEG -37/1.2V TO-220 PKG  
IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG  
IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG  
IC OP AMP LOW-BIAS-H-IMPD 8-DIP-P PKG  
IC MUXR/DATA-SEL CMOS/74HC 2-TO-1-LINE  
IC V RGLTR-V-REF-FXD 1.22/1.24V TO-92  
IC OP AMP LOW-OFS 8-DIP-P PKG  
00000 LT337AT  
3
00000 LT1012CN8  
00000 LT1012CN8  
00000 LT1012CN8  
04713 MC74HC157N  
27014 LM385BZ-1.2  
06665 OP-07CP  
1
1
1
VR1  
W1  
1902-0951  
1
1
DIODE-ZNR 5.1V 5% D0-35 PD=.4W TC .035%  
CABLE ASSEMBLY, SAFETY GROUND  
28480 1902-0951  
01141-61602  
28480 01141-61602  
84  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service  
Replaceable Parts  
Locator Table for Control and Power Supply  
Ref. Grid Ref Grid Ref Grid Ref Grid Ref Grid Ref Grid Ref Grid Ref Grid  
Des. Loc Des Loc Des Loc Des Loc Des Loc Des Loc Des Loc Des Loc  
C1  
C2  
C3  
C4  
C5  
C6  
C7  
C8  
C9  
C-2  
F-3  
H-2  
J-3  
C-3  
C-5  
B-3  
C-5  
C-5  
C17  
C-6 CR10 C-5  
J1  
J2  
J3  
G-7  
B-7  
C-7  
R5  
R6  
R7  
R8  
R9  
B-6  
B-6  
C-4  
C-5  
C-4  
R22 B-3  
R23 B-5  
R24 B-5  
R25 B-2  
R26 B-2  
R27 A-5  
R28 A-5  
R29 B-6  
R30 B-6  
R31 B-3  
R32 B-2  
R40 C-2  
R41 A-2  
R42 C-2  
R43 C-2  
R44 C-3  
R45 C-3  
R46 C-3  
R47 C-2  
R48 D-3  
U5  
U6  
U7  
U8  
U9  
D-6  
B-3  
C-3  
B-3  
A-5  
C18 B-5 CR11 C-3  
C19 B-6 CR12 C-6  
C20 B-5 CR13 C-3  
C21 B-5 CR14 C-6  
C22 B-6 CR17 B-5  
C23 B-4 CR18 B-6  
C24 B-6  
MP1 H-1  
MP2 E-3  
MP3 D-3  
MP4 E-5  
MP7 E-4  
MP8 G-4  
MP9 G-4  
R10 C-5  
R11 C-5  
R12 C-5  
R13 E-3  
R14 C-4  
R15 C-4  
R16 C-4  
R17 A-2  
R18 C-6  
R19 C-6  
R21 B-3  
RP1 B-3  
U10 C-2  
U11 C-3  
S1  
S2  
H-7  
E-2  
C25 B-4 DS1 G-2  
C26 G-6  
VR1 E-3  
C10 C-5  
C11 E-3  
C12 A-3  
C13 A-3  
C14 B-4  
C15 A-3  
C16 C-4  
E1  
F-3 E2  
F-3  
F-6  
E-6  
T1  
H-5  
CR2  
CR3  
P1  
E-7  
U1  
U2  
U3  
U4  
D-4  
D-5  
A-2  
D-3  
CR5 C-3 F1  
CR6 C-5  
E-6  
R3  
R4  
A-6  
A-6  
CR9 C-3 HS1 D-5  
85  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Replaceable Parts  
86  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
Numerics  
100x attenuator test, 32  
10x attenuator test, 30  
cabling diagram, 75  
environmental characteristics, 58  
extension leads, 14  
tests, 27  
A
F
ac adapter, 20  
environmental, 58  
ac coupling adapters, 19  
ac low frequency, 55  
accessories  
features, 2  
general, 58  
performance, 55  
probe connectors, 18  
cleaning instrument, 22  
low response, 44  
function select, 22  
available to order, 8  
circuit connector posts, 15  
extension leads, 14  
ground leads, 14  
mini grabbers, 15  
probe tips, 14  
G
definition, 23  
specification, 55  
test, 36  
shielded signal leads, 15  
supplied with probe, 6  
test board, 26  
grounding, 18  
common mode rejection ratio, 23  
compensating dc, 18  
power & control board, 74  
connector  
using, 13  
accuracy test, attenuator, 30, 32  
adapters, 16  
H
high frequency  
attenuator, 61  
compatibility, 18  
control module, 60  
high-frequency path, 60  
humidity, 58  
attenuator adapter, 49  
initial, 13  
probe, 40, 42  
altitude, 58  
D
amplifiers, 23  
impedance, output, 55  
initial adjustment, 11  
input impedance, 55  
input voltage, maximum, 55  
inspecting, 9  
assemblies, removing & replacing,  
dc gain accuracy, 28  
66  
dc mode, 55  
attenuator  
dc offset, 19, 55  
adapter adjustment, 49  
dc reject, 19, 55  
adapters, 61  
dc reject gain, 11  
differential amplifiers, 23  
differential input range, 55  
differential probe  
B
L
bandwidth  
line voltage, 10  
disassembly, 66  
specification, 55  
test, 34  
blocking dc, 18  
block-level theory, 59  
local control, 61  
reassembly, 67  
locator table, 85  
dimensions, 58  
low frequency response, 44  
direct mail order parts, 71  
low-frequency path, 60  
displayed noises, displayed, 55  
87  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
M
probe gain, 28  
test record, calibration, 27  
testing interval, 27  
theory of operation, 59  
thermal drift, 55  
probe preparation, 41  
making measurements, 13  
manufacturer’s codes, 72  
maximum input voltage, 55  
mini grabbers, 15  
R
troubleshooting  
power module, 64  
N
probe, 64  
remote functions, 22  
noises, 55  
probe control, 64  
remote input connection, 21  
remote offset input, 22  
remote operation, 21  
O
U
offset, 20  
removing assemblies, 66  
using accessories, 13  
offset functions, 61  
replaceable parts, 71  
offset null, 11  
replaceable parts list, 78  
rise time, specification, 55  
operating environment, 58  
operating range, 55  
ordering information, 71  
output impedance, 55  
overload recovery, 55  
variable offset, 22  
vibration, 58  
probe control & power module,  
P
weight, 58  
parts list, 71, 78  
performance specification, 55  
power module, 60  
disassembly, 70  
setup procedure, 11, 12  
troubleshooting, 64  
power requirements, 10, 58  
power supply, 61  
probe adapter  
specifications, performance, 55  
disassembly, 68  
system preparation, 10  
reassembly, 68  
probe adjustment, 40  
probe control  
temperature, 58  
disassembly, 70  
troubleshooting, 64  
probe gain test, 28  
probe tips, 14  
10x attenuator accuracy, 30  
bandwidth, 34  
procedure  
calibration, 27  
attenuator adjustment, 50  
CMRR, 36  
bandwidth, 34  
record calibration, 39  
test board, 18, 26, 62  
test equipment, 9, 26  
CMRR test, 36  
probe adjustment, 42  
88  
Download from Www.Somanuals.com. All Manuals Search And Download.  
commercial license terms, and  
non-DOD Departments and  
Agencies of the U.S. Government  
will receive no greater than  
Restricted Rights as defined in  
FAR 52.227-19(c)(1-2) (June  
1987). U.S. Government users will  
receive no greater than Limited  
Rights as defined in FAR 52.227-  
14 (June 1987) or DFAR 252.227-  
7015 (b)(2) (November 1995), as  
applicable in any technical data.  
Warranty  
Notices  
© Agilent Technologies, Inc.  
2000-2004  
The material contained in this  
document is provided “as is,” and  
is subject to being changed,  
without notice, in future editions.  
Further, to the maximum extent  
permitted by applicable law,  
Agilent disclaims all warranties,  
either express or implied, with  
regard to this manual and any  
information contained herein,  
including but not limited to the  
implied warranties of  
merchantability and fitness for a  
particular purpose. Agilent shall  
not be liable for errors or for  
incidental or consequential  
damages in connection with the  
furnishing, use, or performance  
of this document or of any  
information contained herein.  
Should Agilent and the user have  
a separate written agreement  
with warranty terms covering the  
material in this document that  
conflict with these terms, the  
warranty terms in the separate  
agreement shall control.  
No part of this manual may be  
reproduced in any form or by any  
means (including electronic  
storage and retrieval or  
translation into a foreign  
language) without prior  
agreement and written consent  
from Agilent Technologies, Inc. as  
governed by United States and  
international copyright laws.  
Manual Safety Notices  
CAUTION:  
A CAUTION notice denotes a  
hazard. It calls attention to an  
operating procedure, practice, or  
the like that, if not correctly  
performed or adhered to, could  
result in damage to the product or  
loss of important data. Do not  
proceed beyond a CAUTION  
notice until the indicated  
Manual Part Number  
01141-97002, July 2004  
Print History  
01141-97000, June 2000  
01141-97001, September 2002  
01141-97002, July 2004  
Agilent Technologies, Inc.  
1900 Garden of the Gods Road  
Colorado Springs, CO 80907 USA  
conditions are fully understood  
and met.  
WARNING:  
A WARNING notice denotes a  
hazard. It calls attention to an  
operating procedure, practice, or  
the like that, if not correctly  
performed or adhered to, could  
result in personal injury or death.  
Do not proceed beyond a  
WARNING notice until the  
indicated conditions are fully  
understood and met.  
Technology Licenses  
The hardware and/or software  
described in this document are  
furnished under a license and  
may be used or copied only in  
accordance with the terms of  
such license.  
Restricted Rights  
Legend  
If software is for use in the  
performance of a U.S.  
Government prime contract or  
subcontract, Software is delivered  
and licensed as “Commercial  
computer software” as defined in  
DFAR 252.227-7014 (June 1995),  
or as a “commercial item” as  
defined in FAR 2.101(a) or as  
“Restricted computer software”  
as defined in FAR 52.227-19  
(June 1987) or any equivalent  
agency regulation or contract  
clause. Use, duplication or  
disclosure of Software is subject  
to Agilent Technologies’ standard  
Agilent Technologies  
P.O. Box 2197  
1900 Garden of the Gods Road  
Download from Www.Somanuals.com. All Manuals Search And Download.  
any unintended operation.  
Safety Symbols  
Product Safety  
Notices  
This apparatus has been designed  
and tested in accordance with  
IEC Publication 1010, Safety  
Requirements for Measuring  
Apparatus, and has been supplied  
in a safe condition. This is a  
Safety Class I instrument  
(provided with terminal for  
protective earthing). Before  
applying power, verify that the  
correct safety precautions are  
taken (see the following  
Service instructions are for  
trained service personnel. To  
avoid dangerous electric shock,  
do not perform any service unless  
qualified to do so. Do not attempt  
internal service or adjustment  
unless another person, capable of  
rendering first aid and  
!
Instruction manual symbol: the  
product is marked with this  
symbol when it is necessary for  
you to refer to the instruction  
manual in order to protect  
against damage to the product.  
resuscitation, is present.  
Do not install substitute parts or  
perform any unauthorized  
modification to the instrument.  
Hazardous voltage symbol.  
Capacitors inside the instrument  
may retain a charge even if the  
instrument is disconnected from  
its source of supply.  
warnings). In addition, note the  
external markings on the  
instrument that are described  
under "Safety Symbols."  
Earth terminal symbol: Used to  
indicate a circuit common  
connected to grounded chassis.  
Do not operate the instrument in  
the presence of flammable gasses  
or fumes. Operation of any  
electrical instrument in such an  
environment constitutes a definite  
safety hazard.  
Warnings  
Before turning on the instrument,  
you must connect the protective  
earth terminal of the instrument  
to the protective conductor of the  
(mains) power cord. The mains  
plug shall only be inserted in a  
socket outlet provided with a  
protective earth contact. You  
must not negate the protective  
action by using an extension cord  
(power cable) without a  
Do not use the instrument in a  
manner not specified by the  
manufacturer.  
To clean the instrument  
If the instrument requires  
cleaning: (1) Remove power from  
the instrument. (2) Clean the  
external surfaces of the  
protective conductor  
(grounding). Grounding one  
conductor of a two-conductor  
outlet is not sufficient protection.  
instrument with a soft cloth  
dampened with a mixture of mild  
detergent and water. (3) Make  
sure that the instrument is  
completely dry before  
Only fuses with the required rated  
current, voltage, and specified  
type (normal blow, time delay,  
etc.) should be used. Do not use  
repaired fuses or short-circuited  
fuseholders. To do so could cause  
a shock or fire hazard.  
reconnecting it to a power source.  
If you energize this instrument by  
an auto transformer (for voltage  
reduction or mains isolation), the  
common terminal must be  
connected to the earth terminal of  
the power source.  
Whenever it is likely that the  
ground protection is impaired,  
you must make the instrument  
inoperative and secure it against  
Download from Www.Somanuals.com. All Manuals Search And Download.  

ACP EP Memory Computer Drive EPSNOW20 1GB B User Manual
AGFA Digital Camera DC 8338i User Manual
Agri Fab Lawn Aerator 45 0298 User Manual
Airflow Systems Carbon Monoxide Alarm IAQ50 User Manual
Airstream Automobile INTERNATIONAL CCD User Manual
AKG Acoustics Microphone ULS SERIES User Manual
Altec Lansing MP3 Docking Station Portable Speaker Docking Station User Manual
Audiovox Remote Starter AS 9256 User Manual
BC Speakers Portable Speaker DE 950 User Manual
Behringer DJ Equipment PPA2000BT PPA500BT User Manual