DC MODULE CONNECTORS
Installation Guide
Standard Connector (P/N 5060-3386)
Relay Connector (P/N 5060-3387)
+Sense used to connect the + remote sense lead. Set the
sense switch to Remote when this connection is
used.
General Information
Load connections are made on the connector supplied
with each module. The connector is fastened to the back
of the mainframe. Modules can then be installed or
removed from the mainframe without disturbing the
wiring on the connector.
-Sense used to connect the - remote sense lead. Set the
sense switch to Remote when this connection is
used.
GND
used for signal ground (GND) connections.
The connector has a switch to select either local or remote
voltage sensing, depending on how the load is connected.
Connect the Wires
A relay connector is identified by a check mark on the
Relay label on the connector. Relay connectors include
load and sense disconnect and polarity reversal relays.
Loosen the barrier block screws and connect the wires.
The barrier block accepts spade or ring terminals. You
can also connect a stripped wire to the barrier block.
Twist the wire around under the screw terminal and
tighten the screw.
Making Load Connections
Load connections are the same on both the standard and
the relay connector.
Open the Connector
Open the connector by loosening the four corner screws
(M3 X 10 mm).
TIE
WRAP
Route the wires out of the bottom of the connector as
shown in the figure above. Use the tie-wrap supplied
with the connector to secure the wires to the connector.
CAUTION
• Hand tighten only.
• Start all screws
before tightening.
Close the Connector
• Tighten screws evenly.
• DO NOT overtighten screws.
• DO NOT use locking-type
Close the connector and tighten the four corner screws.
Observe the
on the figure.
CAUTION
Attach the Connector to the Mainframe
A five-terminal barrier block is provided for wire
connections. The barrier block screws are 6-32 x 1/4 inch
(equivalent p/n 2470-0001). The barrier block terminals
are labeled as follows:
Attach the connector to the back of the module on the
mainframe at the appropriate slot location. Tighten the
two screws (M3 x 10mm) on the connector. Observe the
on the figure.
CAUTION
+Output used to connect the + output load wire.
-Output used to conect the - output load wire.
Agilent p/n 5959-3366
May, 2000
Microfiche p/n 5959-3367