IDEAL INDUSTRIES INC.
TECHNICAL MANUAL
MODEL: 61-362
The Service Information provides the following information
• Precautions and safety information
• Specifications
• Performance test procedure
• Calibration and calibration adjustment procedure
• Basic maintenance (replacing the battery and fuses)
Form number: 61362
Rev 1. date: May 2001
Form number 61362
rev 1 May 2002
Download from Www.Somanuals.com. All Manuals Search And Download.
Page 1
Introduction
Warning
To avoid shock or injury, do not perform the verification tests or calibration
procedures described in this manual unless you are qualified to do so.
The information provided in this document is for the use of qualified personnel
only.
Caution
The 61-362 serials contain parts that can be damaged by static discharge.
Follow the standard practices for handling static sensitive devices.
For additional information about IDEAL INDUSTRIES and its products,
and services, visit IDEAL INDUSTRIES web site at:
SAFETY
Review the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. To avoid potential hazards, use the product only as specified.
It is recommended that you read through the Operation or User manual before starting. Not all Caution,
Warning, or Danger precautions are listed in this manual.
CAUTION.
These statements identify conditions or practices that could result in damage to the equipment or other
property.
WARNING.
These statements identify conditions or practices that could result in personal injury or loss of life.
Specific precautions
Use proper Fuse. To avoid fire hazard, use only the fuse type and rating specified for this product.
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the
product without the covers in place.
Electric overload. Never apply a voltage to a connector on the product that is outside the range specified
for that connector.
Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads
while they are connected to a voltage source.
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or
damp conditions.
Form number 61362
rev 1 May 2002
Download from Www.Somanuals.com. All Manuals Search And Download.
Page 2
General specifications
Model 61-362
Characteristics
Display
Description
3 ½ Digit LCD display
3200 count, maximum reading 3199
32 segment analog bar graph
“OL” displayed
Display Count
Over range Indication
Sampling Rate
2 time/second
Operating Environment:
Relative Humidity
Temperature Coefficient
Storage Environment:
Power source:
0°C to 50°C (32°F to 122°F) 70%RH
0.05X x (accuracy) per °F (32°F to 65°F, 85°F to 122°F
-20°C to 60°C (0°F to 140°F) at <70 relative humidity
9V Battery (NEDA 1604)
Battery Live:
200 hours typical (alkaline)
Low Battery Indicator:
Auto Power Off mode
A protection Fuse
symbol indicates low battery voltage
Approximately 25 minutes
0.5A/250V fast acting fuse Type LA-3895
10A/600V fast acting fuse, Type LA-3897
Dimensions
Weight:
Safety
5.75” H X 2.75” W X 1.5” D {without holster}
Approximately 18.0 oz. including battery
UL1244, and Design to comply with IEC 1010-1 Cat III
RANGES and ACCURACY SPECIFICATION
61-362
Function Setting
Ranges
Accuracy
3.200V/32.00V/320.0V/ 600V
50Hz to 500Hz
2.0% ± 4 digits
AC Voltage
320.0mV/32.00V/320.0V/600V
3.200V
320µA/3200 µA/32.00mA/320mA
10A,
50Hz to 500Hz
1.2% ± 1digit
0.8% ± 1digit
2.5% ± 4 digits
3.5% ± 4 digits
DC Voltage
AC Current
DC Current
320µA/3.20mA/32.00mA
2.0% ± 1 digit
10A,
320.0Ω
3.200k/32.00k/320.0k
3.200MΩ
32.00MΩ
beep on < 100Ω
2V DC max
3.0% ± 1 digit
2.0% ± 3 digits
1.5% ± 3 digits
2.5% ± 3 digits
5.0% ± 5 digits
Not Specified
10% ± 2 digit
Resistance
Continuity
Diode Check
AC Converter:
Overload Protection:
Average responding, RMS Calibrated to Sine Wave
AC and DC Volts: 600V DC or AC rms .
Resistance, Diode, Continuity: 500V DC or AC rms
mA input: .5A/250V DC/AC rms
Amps input: 10A/600V DC/AC rms
Form number 61362
rev 1 May 2002
Download from Www.Somanuals.com. All Manuals Search And Download.
Page 3
PERFORMANCE VERIFICAITONS
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning
correctly. If the meter does not conform to any of the listed limits the calibration procedure must be
performed.
Performance Verification Preparation
1. Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization
should be reached after 30 minutes.
2. Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5 V
DC. If the battery measures under 7.5V DC, replace the battery before beginning the performance test.
3. Input the values listed in Table 1 for the Standard 61-362
Table 2 Performance Verification 61-362
Function Setting /Range
ACV 3.2V
ACV 3.2V
ACV 32
ACV 32
ACV 320
ACV 320
ACV 600
ACV 600
DCV 320mV
DCV 3.2V
DCV 32V
DCV 320
DCV 600
Input
Low Limit
2.936
2.936
29.36
29.36
293.6
293.6
486
High Limit
3.064
3.064
30.64
30.64
306.4
306.4
514
3.000V @ 50Hz
3.000V @ 500Hz
30.00 @ 50Hz
30.00 @ 500Hz
300V AC @ 50Hz
300V AC @ 500Hz
500V AC @ 50Hz
500V AC@ 60Hz
300mV DC
3.000V DC
30.00V DC
300.0V DC
500V DC
486
514
296.3
2.975
29.63
296.3
493
303.7
3.025
30.37
303.7
507
A DC 320µA
A DC 3200 µA
A DC 32mA
A DC 320mA
A DC 10
300µA
3000µA
30mA
300mA
293.9
2939
29.39
293.9
8.70
306.1
3061
30.61
306.1
9.30
9.00 DCA
A AC 320µA
A AC 320µA
A AC 3200µA
A AC 320oµA
A AC 32mA
A AC 32mA
A AC 320mA
A AC 320mA
A AC 10A
A AC 10A
Ω 320
Ω 2k
Ω 32k
Ω 320k
Ω 2M
Ω 32M
Diode Test
Continuity Test
Continuity Test
300µA @ 50Hz
300µA @ 500Hz
3000µA @ 50Hz
3000µA @ 500Hz
30mA @ 50Hz
30mA @ 500Hz
300mA @ 50Hz
300mA @500Hz
9A @ 50Hz
9A @ 500Hz
300.0
3.000k
30.00k
300.0k
3.000M
30.00M
500mV
120Ω
80Ω
292.1
292.1
2921
307.9
307.9
3079
3079
30.79
30.79
307.9
307.9
9.36
2921
29.21
29.21
292.1
292.1
8.64
8.64
9.36
297.7
2.950
29.52
295.2
2.922
28.45
.448
306.3
3.048
30.48
304.8
3.078
31.55
.552
Beep off
Beep on
Form number 61362
rev 1 May 2002
Download from Www.Somanuals.com. All Manuals Search And Download.
Page 4
CALIBRATION
Calibration Preparation
Required Equipment
The class of calibrator or equipment should have an accuracy that exceeds, by an expectable ratio the
accuracy of this instrument.
1. Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization
should be reached after 30 minutes.
2. Disconnect the test leads and turn the range switch to “OFF”.
3. Remove the screw holding the bottom case cover, just above the battery cover.
4
The case bottom is secured to the case top by two internal snaps. (at the LCD end).
lift up on the battery end until the case un-snaps.
5
Using a calibrated meter ensure the battery measures a minimum of 7.5 V dc.
If the battery measures under 7.5V DC, replace the battery.
Calibration Procedure
It is recommended that all IDEAL meters undergo the following calibration procedure on
an annual basis.
61-362 Calibration Procedure.
Calibration
1. Set the Function/Range Switch to the “3.200V DC” position.
2. Set the output of the DC calibrator for 3.000V and connect it to the “V-Ohm” and “COM”
input terminals.
3. Adjust VR1 (VR 200 ohm) until the display reads 3.000V +/- 1 digit.
Note : This is the only adjustments required for the 61-362. Calibration is complete.
Battery Replacement (refer to Figure 2)
1. Disconnect the test leads from any circuit under test and turn off meter.
2.Remove the three screws for the back case cover.
3. Remove battery from compartment noting the “+” and “- “ position of the Battery terminals.
4. Remove the Battery plastic sleeve and place it on the new battery
(Damage can occur to circuit if Plastic sleeve is not replace with new Battery)
5. Install new 9V battery into compartment and assure proper polarity of battery.
(An alkaline type NEDA #1604 is recommended.)
6. Install bottom case cover and secure with screws.
Form number 61362
rev 1 May 2002
Download from Www.Somanuals.com. All Manuals Search And Download.
Page 5
Replacing Fuse (refer to Figure 3)
1. Disconnect the test leads and turn the range switch to “OFF”.
2. Remove the three screws holding the bottom case cover
3
Us a digital multimeter in low resistance {ohms} mode to check the two fuses
mA input, 0.5A / 250V fast acting fuse.
Amp input, 10A / 600V fast acting fuse
4. Remove the defective fuse with the recommended fuse
mA with : 0.5A/250V fast acting fuse Type LA-3895 is recommended.
Amp with: 10A/600V fast acting fuse, Type LA-3897 is recommended
5. Install bottom cover and secure with screws.
Form number 61362
rev 1 May 2002
Download from Www.Somanuals.com. All Manuals Search And Download.
|