Agilent Technologies Computer Hardware 66000A User Manual

INSTALLATION GUIDE  
Agilent Technologies  
MPS Mainframe  
Model 66000A  
SERIAL NUMBERS  
For mainframe with serial numbers:  
Agilent Model 66000A: 3138A-00101 and above  
* For instruments with higher Serial Numbers, a change page may be included.  
Agilent Part No. 66000-90001  
Microfiche Part No. 66000-90002  
Printed in USA: June , 1997  
Update April 2000  
SAFETY SUMMARY  
The following general safety precautions must be observed during all phases of operation, service, and repair of this  
mainframe. Failure to comply with these precautions or with specific warnings elsewhere in this guide violates safety  
standards of design, manufacture, and intended use of the mainframe. Agilent Technologies assumes no liability for the  
customer’s failure to comply with these requirements.  
BEFORE APPLYING POWER.  
Verify that the product is set to match the available line voltage and the correct fuse is installed.  
GROUND THE MAINFRAME.  
This product is a Safety Class 1 instrument (provided with a protective earth terminal). To minimize shock hazard, the mainframe must  
be connected to an electrical ground. The mainframe must be connected to the ac power supply mains through a three-conductor power  
cable, with the third wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective  
(grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal  
injury.  
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE.  
Do not operate the instrument in the presence of flammable gases or fumes.  
KEEP AWAY FROM LIVE CIRCUITS.  
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified  
service personnel. Do not replace components with power cable connected. Under certain conditions, dangerous voltages may exist even  
with the power cable removed. To avoid injuries, always disconnect power, discharge circuits and remove external voltage sources before  
touching components.  
DO NOT SERVICE OR ADJUST ALONE.  
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.  
DO NOT EXCEED INPUT RATINGS.  
This mainframe is equipped with a power line filter to reduce electromagnetic interference and must be connected to a properly grounded  
receptacle to minimize electric shock hazard. Operation at line voltages or frequencies in excess of those stated on the line rating label  
may cause leakage currents in excess of 5.0 mA peak.  
SAFETY SYMBOLS.  
Instruction manual symbol: the product will be marked with this symbol when it is necessary for the user to refer to the  
instruction manual (refer to Table of Contents) .  
Indicates hazardous voltages.  
Indicates earth (ground) terminal.  
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly  
performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the  
indicated conditions are fully understood and met.  
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like, which, if not correctly  
performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed  
beyond a CAUTION sign until the indicated conditions are fully understood and met.  
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT.  
Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the  
instrument. Return the instrument to an Agilent Technologies Sales and Service Office for service and repair to ensure that safety features  
are maintained.  
Mainframes that appear damaged or defective should be made inoperative and secured against unintended operation until they can be  
repaired by qualified service personnel.  
3
SAFETY SUMMARY (continued)  
GENERAL: Any LEDs used in this product are Class 1 LEDs as per IEC 825-1.  
ENVIRONMENTAL CONDITIONS:  
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is  
designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the  
specifications tables for the ac mains voltage requirements and ambient operating temperature range.  
SAFETY SYMBOL DEFINITIONS  
Symbol  
Description  
Direct current  
Symbol  
Description  
Terminal for Line conductor on permanently  
installed equipment  
Alternating current  
Caution, risk of electric shock  
Both direct and alternating current  
Three-phase alternating current  
Earth (ground) terminal  
Caution, hot surface  
Caution (refer to accompanying documents)  
In position of a bi-stable push control  
Protective earth (ground) terminal  
Frame or chassis terminal  
Out position of a bi-stable push control  
On (supply)  
Terminal for Neutral conductor on  
permanently installed equipment  
Off (supply)  
Terminal is at earth potential  
Standby (supply) - Units with this symbol are not  
completely disconnected from ac mains when  
this switch is off. To completely disconnect the  
unit from ac mains, either disconnect the power  
cord or have a qualified electrician install an  
external switch.  
(Used for measurement and control  
circuits designed to be operated with  
one terminal at earth potential.)  
Herstellerbescheinigung  
Diese Information steht im Zusammenhang mit den Anforderungen der  
Maschinenläminformationsverordnung vom 18 Januar 1991.  
* Schalldruckpegel Lp <70 dB(A) * Am Arbeitsplatz * Normaler Betrieb  
* Nach EN 27779 (Typprüfung).  
Manufacturer’s Declaration  
This statement is provided to comply with the requirements of the German Sound Emission Directive,  
from 18 January 1991.  
* Sound Pressure Lp <70 dB(A) * At Operator Position * Normal Operation  
* According to EN 27779 (Type Test).  
4
DECLARATION OF CONFORMITY  
according to ISO/IEC Guide 22 and EN 45014  
Manufacturer's Name:  
Manufacturer's Address:  
Agilent Technologies, Inc.  
140 Green Pond Road  
Rockaway, New Jersey 07866  
U.S.A.  
declares that the Product  
Product Name:  
Modular Power System mainframe and modules  
Model Number:  
Agilent 66000A mainframe and modules  
Agilent 66101A, 66102A, 66103A, 66104A, 66105A, 66106A  
conforms to the following Product Specifications:  
Safety:  
EMC:  
IEC 348:1978 / HD 401S1: 19811  
CISPR 11:1990 / EN 55011:1991 - Group 1 Class A  
IEC 801-2:1991 / EN 50082-1:1992 - 4 kV CD, 8 kV AD  
IEC 801-3:1984 / EN 50082-1:1992 - 3 V / m  
IEC 801-4:1988 / EN 50082-1:1992 - 0.5 kV Signal Lines  
1 kV Power Lines  
Supplementary Information:  
The product herewith complies with the requirements of the Low Voltage Directive  
73/23/EEC and the EMC Directive 89/336/EEC and carries the CE-marking accordingly.  
Note 1: The product family was introduced prior to 12/93.  
New Jersey  
Location  
January 1997  
Date  
Bruce Krueger / Quality Manager  
European Contact: Your local Agilent Technologies Sales and Service Office or Agilent Technologies GmbH,  
Department TRE, Herrenberger Strasse 130, D-71034 Boeblingen (FAX:+49-7031-14-3143)  
Printing History  
The current edition of this guide is indicated below. Reprints of this guide containing minor corrections and updates may  
have the same printing date. New editions are identified by a new printing date and, in some cases, by a new part number.  
A new edition incorporates all new or corrected material since the previous edition. Changes to the guide occurring  
between editions are covered by change sheets shipped with the guide. Also, if the serial number prefix of your mainframe  
is higher than those listed on the title page of this guide, then it may or may not include a change sheet. That is because  
even though the higher serial prefix indicates a design change, that change may not affect the content of the guide.  
Edition 1 ... October, 1991 (Agilent Model 66000A)  
Update April 2000  
© Copyright 1991 Agilent Technologies, Inc.  
This document contains proprietary information protected by copyright. All rights are reserved. No part of this document  
may be photocopied, reproduced, or translated into another language without the prior consent of Agilent Technologies.  
The information contained in this document is subject to change without notice.  
5
Table Of Contents  
1.  
2.  
General Information  
Introduction ..................................................................................................................................................9  
Safety Considerations...................................................................................................................................9  
Instrument Identification ..............................................................................................................................9  
Options and Parts .......................................................................................................................................10  
Description .................................................................................................................................................10  
Specifications and Supplemental Characteristics .......................................................................................11  
Installation  
Installation..................................................................................................................................................13  
Items Supplied............................................................................................................................................13  
Input Power ................................................................................................................................................13  
Line Voltage Conversion.........................................................................................................................13  
Fuses........................................................................................................................................................14  
Power Cord Installation..............................................................................................................................14  
Location and Cooling .................................................................................................................................15  
Bench Operation......................................................................................................................................15  
Rack Mounting........................................................................................................................................15  
Temperature Performance .......................................................................................................................15  
Rear Panel Switches ...................................................................................................................................16  
Address Switches .................................................................................................................................16  
Frame Mode Switch ................................................................................................................................16  
Module Installation.....................................................................................................................................18  
Module Removal........................................................................................................................................18  
Mainframe Turn-On Checkout ...................................................................................................................19  
3.  
Connections  
Rear Panel ..................................................................................................................................................21  
Module Connections...................................................................................................................................21  
Controller Connections...............................................................................................................................22  
Multiple MPS Mainframes......................................................................................................................22  
MPS Mainframe/Power Supply Connections..........................................................................................22  
Keyboard Connection.................................................................................................................................22  
Digital Connections....................................................................................................................................22  
Trigger Connections...................................................................................................................................24  
Index  
..............................................................................................................................27  
6
Figures  
2-1.  
2-2.  
2-3.  
2-4.  
2-5.  
3-1.  
3-2.  
3-3.  
3-4.  
Power Cord Plug Configurations............................................................................................14  
Power Cord Installation..........................................................................................................15  
Rear Panel Switches ...............................................................................................................16  
Switch Settings.......................................................................................................................17  
Module Pull Tab.....................................................................................................................18  
Rear Panel Connections..........................................................................................................21  
Controller Connections...........................................................................................................23  
Digital Connector ...................................................................................................................24  
Examples of the Remote Inhibit Function ..............................................................................25  
Tables  
1-1.  
Specifications and Supplemental Characteristics............................................................................11  
7
1
General Information  
Introduction  
This guide describes how to install the Agilent 66000A Modular Power System (MPS) Mainframe. It includes the  
following information:  
specifications and operating characteristics  
connecting the power cord to the mainframe  
setting the address and frame switches  
installing modules in the mainframe  
mainframe rear panel connections  
Mainframe turn-on and checkout procedures are included with the module manual since an installed module is required to  
properly check out mainframe operation. Refer to the module User’s Guidefor more information.  
Related Documents  
Programming Guide for Agilent Series 66lxxA MPS Power Modules  
Installation Guide for DC Module Connectors  
Agilent Part No.  
5959-3362  
5959-3366  
User’s Guide for Agilent 66lxxA Power Modules  
5959-3360  
Service Guide for Agilent Series 66lxxA MPS Power Modules  
Service Guide for Agilent 66000A Modular Power System Mainframe  
Quick Reference Card for optional Agilent 66001A MPS Keyboard  
Agilent 66000A Modular Power System Product Note  
5959-3364  
66000-90003  
66001-90001  
5091-2497E  
Safety Considerations  
This mainframe is a Safety Class 1 instrument, which means that it has a protective earth terminal. That terminal must be  
connected to earth ground through a power source equipped with a 3-wire ground receptacle.  
Refer to the safety summary page at the beginning of this manual for general safety information. Before installation or  
operation, check the mainframe and review this manual for safety markings and instructions. Safety information for specific  
procedures is located at the appropriate places in this manual.  
Instrument Identification  
The mainframe is identified by a two-part serial number, such as 3140A-01234, on the back of the unit. The first part, or  
prefix, is a number-letter combination that describes:  
3140 The year and week of manufacture or last significant design change. Add 1960 to the first two  
digits to determine the year. For example, 31=1991, 32=1992, etc. The last two digits specify the week  
of the year (40 = the 40th week).  
A
The letter indicates the country of manufacture, where A = USA.  
9
General Information  
Options and Parts  
The following options are available with the mainframe:  
Option  
Description  
908  
909  
910  
Rack Mounting Kit  
Rack Mounting Kit with handles  
Extra Manual Set  
Power Cords  
831  
833  
834  
841  
843  
845  
846  
847  
848  
12 AWG unterminated power cord  
1.5 mm2 unterminated power cord  
10 AWG unterminated power cord  
20 A/250 V NEMA 6-20P plug with 12 AWG power cord  
20 A/250 V JIS C8303 plug with 12 AWG power cord  
16 A/220 V IEC 309 plug with 1.5 mm2 power cord  
30 A/120 V L5-30P locking plug with 10 AWG power cord  
16 A/220 V CEE 7/7 plug with 1.5 mm2 power cord  
15 A/240 V BS 546 plug with 1.5 mm2 power cord  
The following mainframe items can be ordered separately:  
Part Description  
Safety cover, ac input  
Strain relief  
Part Number  
5040-1676  
1400-1195  
0515-2086  
0360-2421  
2470-0001  
5080-2148  
1252-1488  
66000-40006  
66000-90001  
Screw, strain relief  
Power barrier block (Figure 2-2)  
Screw, power barrier block  
Serial link cable (1 meter)  
Digital connector plug (Figure 3-3)  
Filler panel (8 supplied)  
Mainframe Installation Guide  
Description  
The Agilent 66000 Modular Power System (MPS) is an integrated, fully programmable power system that saves rack space  
and GPIB addresses, and is easy to install and use. Each Agilent 66000A MPS Mainframe can accommodate up to eight  
single-slot modules. With an auxiliary mainframe connected through a serial link cable, up to sixteen single-slot modules  
can be controlled from a single GPIB address. For applications requiring a broader range of power ratings, the Agilent  
66000 system can be serial-linked to the Agilent 667xA, 665xA, 664xA, and 603xA series power supplies.  
Modules are removed and installed through the front of the mainframe. Each module has a connector that attaches to the  
back of the mainframe. With the module connector wired and attached to the mainframe, the module slides in and out of the  
mainframe without disturbing the application connections.  
The optional Agilent 66001A MPS Keyboard is available for controlling the Agilent 66000 Modular Power System without  
having to connect it to a computer.  
10  
General Information  
Specifications and Supplemental Characteristics  
Table 1-1 lists the specifications and supplemental characteristics of the MPS mainframe. Module specifications are  
included in the appropriate module Users Guide. Specifications are warranted in the temperature range of 0 to 55 degrees  
C. Supplemental characteristics are not warranted, but are descriptions of minimum, maximum, or average performance  
that has been determined by design or type testing.  
Table 1-1. Specifications and Supplemental Characteristics  
PERFORMANCE SPECIFICATIONS  
AC Input Ratings  
Standard:  
87- 132 Vac  
174- 250 Vac  
Frequency tolerance:  
47 - 63 Hz  
SUPPLEMENTAL CHARACTERISTICS  
Maximum AC Line Current Ratings (with 8 power supply modules operating at maximum output)  
100 Vac:  
29 A rms  
120 Vac:  
25 A rms  
200 Vac:  
16 A rms  
220/230 Vac:  
240 Vac:  
16 A rms  
15 A rms  
Maximum power input:  
3200 VA; 1800 W maximum; 1600 W typical  
GPIB Interface Capabilities  
Languages:  
Standard Commands for Programmable Instruments (see module  
Programming Guide)  
IEEE 488.2 Capabilities:  
SH1, AH1, T6, L4, SR1, RL1, PPO, DC1, DT1, El, and CO  
Serial Link Capabilities (Multiple mainframes sharing one GPIB address)  
Number of mainframes:  
Number of modules:  
2
16  
Serial-linked power supplies:  
total number of modules and serial-linked power supplies (Agilent  
667xA/665xA/664xA/603xA) cannot exceed 16  
30 m (100 ft) maximum  
Chain length:  
TrigIn/TrigOut Characteristics  
Maximum Ratings:  
± 16.5 V between center conductor and chassis  
Trig Out (HC TTL output):  
V OL @ 4 ma = 1.34 V maximum  
VOL @ 2 ma = 0.8 V maximum  
VOH 4 ma = 2.66 V minimum  
VOH 2 ma = 3.18 V minimum  
Trig In (10k pull-up):  
I IL @ 4 V = 700 µA  
I
IH @ 5.25 V = 200 µA  
V IL = 0.8 V maximum  
V IH = 2.0 V minimum  
11  
General Information  
Table 1-1. Specifications and Supplemental Characteristics (continued)  
INH/FLT Characteristics  
Maximum Ratings:  
± 16.5 V between pins 3 and 4  
+ 16.5 V between pins 1 and 2  
- 5.3 V between pins 1 and 2  
INH (10k pull-up):  
I IL @ 4 V = 700 µA  
I IH @ 5.25 V = 200 µA  
VIL = 0.8 V maximum  
VIH = 2.0 V minimum  
DFI (open collector):  
V CE = 0.4 V maximum @ IC = 2.2 mA  
I IH @ 5.25 V = 200 µA  
VIL = 0.8 V maximum  
VIH = 2.0 V minimum  
Dimensions  
Width:  
Height:  
Depth:  
426 mm (16.75 in)  
178 mm (7 in), add 13 mm (0.5 in) for removable feet  
678 mm (26.7 in), including module connectors  
Weight (does not include modules)  
Net:  
Shipping:  
15 kg/33 lbs.  
19 kg/41 lbs.  
Environmental & Safety  
Compliance  
CSA 556B IEC 348 UL 1244 VDE 0411  
VDE 0871.6.78 Level B  
RFI Suppression  
12  
General Information  
2
Installation  
Installation  
When you receive your mainframe, first look for any obvious damage that could have occurred during shipment. If there is  
damage, notify the carrier and the nearest Agilent Sales Office immediately. Warranty information is printed behind the  
cover page of this manual.  
Save the shipping carton and packing materials in case the mainframe has to be returned to Agilent Technologies in the  
future. If you return the mainframe for service, attach a tag identifying the owner and model number. Also include a brief  
description of the problem.  
Items Supplied  
In addition to this manual, check that the following items are included with the mainframe:  
Power Cord  
Your mainframe was shipped with a power cord for the type of outlet used at your location. Depending  
on the option that was ordered, some power cords are shipped without a plug. If the appropriate cord was not included,  
refer to the option list in Chapter 1 and contact your nearest Agilent Sales and Service Office to obtain the correct cord.  
Your mainframe cannot use a standard power cord. The power cords listed in Chapter 1 have heavier  
gauge wire.  
Filler Panels  
Serial Cable  
Eight filler panels are included with each mainframe. Filler panels are installed in empty module slots.  
A 1-meter serial cable that connects to the J1 connector next to the GPIB connector. This is used to  
serially connect two mainframes or a mainframe with power supplies that have the serial-link capability on the same GPIB  
address.  
Keyboard  
Reference  
Card  
A quick reference card for the optional Agilent 66001A MPS Keyboard is included with the  
mainframe.  
Manual  
If applicable, change sheets may be included with this manual. If change sheets are included, make  
Change Sheet the corrections to the manual.  
Input Power  
You can operate your mainframe either from a 87 - 132 Vac or a 174 - 250 Vac single phase ac power source. Refer to  
Chapter 1 for complete input power specifications and requirements.  
Line Voltage Conversion  
The mainframe contains no line voltage switches. Refer to the module User’s Guidefor information on changing the line  
voltage.  
13  
Installation  
Fuses  
The mainframe fuse is located inside the mainframe. Refer to the mainframe Service Manual for information about  
replacing the fuse.  
Power Cord Installation  
The power cord supplied with your mainframe may or may not include a power plug. Figure 2-1 shows the various power  
plugs. Ring terminals are attached to one end of the power cord to facilitate mainframe installation.  
Figure 2-1. Power Cord Plug Configurations  
If your power cord will be directly wired into your location’s power system, installation must be done  
by a qualified electrician and in accordance with local electrical codes.  
To install the power cord, refer to Figure 2-2 and proceed as follows:  
Position the strain relief connector, safety cover, rubber boot, and connector nut on the power cord.  
Secure the ground wire to the earth ground stud.  
Connect the neutral wire to the "N" input terminal.  
Connect the line wire to the "L" input terminal.  
Note  
Route the power cord wires upward from the terminal block. This allows room to install the safety cover.  
Replace the safety cover and tighten the cover and strain relief screws.  
Do not apply power to the mainframe until directed to do so in the Check-out procedure located in the  
module User’s Guide.  
14  
Installation  
Figure 2-2. Power Cord Installation  
Location and Cooling  
Bench Operation  
Chapter 1 gives the dimensions of your mainframe. The mainframe has plastic feet that are shaped to ensure self-alignment  
when stacked with other Agilent System II cabinets. The feet may be removed for rack mounting. Your mainframe must be  
installed in a location that allows sufficient space at the sides and rear of the mainframe for adequate air circulation. Do not  
block the vents at the sides and rear of the mainframe.  
Rack Mounting  
The mainframe can be mounted in a standard 19-inch rack panel or cabinet. Refer to Chapter 1 for mainframe dimensions.  
A fully loaded mainframe can weigh over 36 kg (80 lbs). Install the mainframe into the rack before installing the modules  
into the mainframe.  
Rack mounting kits are available as option numbers 908 and 909 (with handles). Installation instructions are included with  
each rack mounting kit.  
Note  
Instrument support rails are also required for rack mounting. These are normally supplied with the  
cabinet and are not included with the rack mounting kits.  
Temperature Performance  
The mainframe can operate within the temperature range of 0° to 55° C. Fans in each module cool the modules by drawing  
in air through the sides of the mainframe and exhausting it out the back. Using Agilent rack mount or slide kits will not  
impede the flow of air.  
15  
Installation  
Rear Panel Switches  
The six-position rear panel switch is shown in the following figure.  
Figure 2-3. Rear Panel Switches  
Address Switches  
Address switches A0 through A4 set the primary GPIB address. The address is set in binary, A0 being the least significant  
bit and A4 being the most significant bit. Figure 2-3 shows the factory-set address of "5" (binary 00101). Any address from  
decimal 0 to decimal 30 is a valid GPIB address. Refer to the following diagram for switch settings.  
Address  
Switch Positions  
Address  
Switch Positions  
Address  
Switch Positions  
A4 A3 A2 A1 A0  
A4 A3 A2 A1 A0  
A4 A3 A2 A1 A0  
0
1
2
3
4
5
6
7
8
9
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
1
1
0
0
0
0
1
1
1
1
0
0
0
0
1
1
0
0
1
1
0
0
0
1
0
1
0
1
0
1
0
1
10  
11  
12  
13  
14  
15  
16  
17  
18  
19  
0
0
0
0
0
0
1
1
1
1
1
1
1
1
1
1
0
0
0
0
0
0
1
1
1
1
0
0
0
0
1
1
0
0
1
1
0
0
1
1
0
1
0
1
0
1
0
1
0
1
20  
21  
22  
23  
24  
25  
26  
27  
28  
29  
30  
1
1
1
1
1
1
1
1
1
1
1
0
0
0
0
1
1
1
1
1
1
1
1
1
1
1
0
0
0
0
1
1
1
0
0
1
1
0
0
1
1
0
0
1
0
1
0
1
0
1
0
1
0
1
0
Frame Mode Switch  
The frame mode switch determines the IEEE 488 secondary addresses (also referred to as subaddresses) of the modules that  
are installed in the mainframe. The frame mode switch must be set to MAIN when the mainframe is connected directly to  
the GPIB. With the frame mode switch set to MAIN, modules are assigned secondary addresses of 0 to 7, which  
corresponds to the slot locations indicated on the mainframe.  
The frame mode switch must be set to AUX (auxiliary) when the mainframe is serially-linked to another mainframe that is  
directly connected to the GPIB. This also applies when the mainframe is serially linked to an Agilent 603xA/664xA/665xA/  
667xA power supply that is directly connected to the GPIB.  
In mainframes that have the frame mode switch set to AUX, modules are assigned secondary addresses of 8 to 15.  
Secondary address 8 corresponds to slot location 0 on the mainframe and secondary address 15 corresponds to slot location  
7 on the mainframe as shown in the following diagram.  
16  
Installation  
Mainframe Slot Locations/Addresses  
Mainframe Mode  
Mainframe Mode  
Switch = AUX  
Switch = MAIN  
Slot Location  
Secondary Address  
0
0
1
1
2
2
3
3
4
4
5
5
6
6
7
7
0
8
1
9
2
3
4
5
6
7
10 11 12 13 14 15  
Note  
Secondary addresses are reserved for the mainframe whether modules occupy the associated slot locations  
or not.  
You may want to apply stick-on labels over the slot locations of the mainframe to indicate that the modules installed in the  
auxiliary mainframe are at secondary addresses 8 to 15. You may also label the modules according to their slot locations.  
If any modules are removed, they can then be reinstalled in the same slot.  
Figure 2-4 illustrates the relationship of the switch settings and secondary addresses. Refer to Chapter 3 for connection  
information.  
Figure 2-4. Switch Settings  
17  
Installation  
Module Installation  
Note  
A fully loaded mainframe can weigh over 36 kg (80 lbs). Install the mainframe into the rack before  
installing the modules into the mainframe.  
All modules install through the front of the mainframe. Installation (and removal) can be accomplished even when the  
mainframe is rack-mounted. To install the modules in the mainframe, refer to Figure 2-5 and proceed as follows:  
Use a flat-bladed screwdriver and release the pull tab by pressing down.  
Extend the pull tab from the module so that the serial number is visible.  
Slide the module into the slot until it is fully inserted in the mainframe.  
Push the pull tab into the front of the module. The slot number should be visible below the pull tab.  
Note  
If the module is not fully inserted in the mainframe, the pull tab will not retract into the module.  
Install filler panels in the front of any unused mainframe slots.  
Install the module connectors on the back of the mainframe.  
Note  
Instructions for wiring and installing module connectors are included with each module.  
Figure 2-5. Module Pull Tab  
Module Removal  
All modules are removed through the front of the mainframe. Modules may be removed even when the mainframe is turned  
on.  
18  
Installation  
There must be no or minimal current flowing through the connector pins at the back of the module  
that is being removed. This can be accomplished by disabling the module. If this precaution is  
repeatedly violated at high current, the connector pins at the back of the module  
will be damaged.  
There must be no activity on the GPIB bus, the trigger connectors, or the FLT/INH connector of  
the mainframe when a module is removed. This can be accomplished by not generating any GPIB  
or FLT/INH activity, and by setting all modules to the trigger idle state (refer to the module  
Programming Guide). If this precaution is violated, GPIB commands to other modules may be  
corrupted, trigger or FLT/INH data may be lost, or an inadvertent trigger may be generated in  
another module.  
To remove the module from the mainframe, proceed as follows:  
Use a flat-bladed screwdriver and release the pull tab by pressing down (see Figure 2-5).  
Grasp the pull tab and pull the module out of the mainframe.  
Note  
The module connector remains attached to the mainframe after the module has been removed.  
Mainframe Turn-On Checkout  
The turn-on checkout procedures that verify module operation also verify the operation of the MPS mainframe. You cannot  
verify the operation of the mainframe without having at least one installed module. The module self-test sequence checks  
the following mainframe functions:  
mainframe microprocessor  
GPIB interface  
module/mainframe communications link  
Refer to the module User’s Guidefor complete information on the turn-on checkout procedure.  
19  
Installation  
3
Connections  
Rear Panel  
Figure 3-1 shows where all of the rear panel connections to the mainframe are made.  
Figure 3-1. Rear Panel Connections  
Module Connections  
Wire connections to the modules are made using a separate module connector that is supplied with each module. After the  
module connector has been wired, it simply attaches to the back of the mainframe. You can attach the module connector to  
the mainframe either before or after the module is installed. Instructions for wiring and attaching the connector are provided  
with each module.  
Note  
Install the module connectors from left to right on the back of the mainframe. This makes it easier to  
access the screws that attach the connectors to the mainframe.  
21  
Connections  
Controller Connections  
If you have only one MPS mainframe in your system, simply connect it to the controller using the GPIB cable. If you have  
more than one mainframe, or if you also have an Agilent 667xA, 665xA, 664xA, or 603xA power supply in your system,  
you can use the serial-link cable supplied with your mainframe to facilitate connections. Figure 3-2 illustrates the serial-link  
connections.  
Multiple MPS Mainframes  
See Figure 3-2. Up to 16 modules may be controlled from a primary GPIB bus address by linking two mainframes.  
The first mainframe in a linked configuration is directly connected to the controller via a GPIB cable. The frame mode  
switch on the back of this mainframe is set to "MAIN". The first mainframe has a unique primary GPIB address and  
the modules in this mainframe use secondary addresses (also referred to as subaddresses) 0 through 7.  
The second or "linked" mainframe, connects to the first mainframe with the serial-link cable. The frame mode switch  
on the back of this mainframe is set to "AUX". The second mainframe uses the same primary address as the first  
mainframe, but the modules in the second mainframe use secondary addresses 8 through 15.  
MPS Mainframe/Power Supply Connections  
See Figure 3-2b. Only one mainframe can be serially linked to a system that includes Agilent 667xA, 665xA, 664xA, or  
603xA power supplies. Because the mainframe has only one serial-link jack, it can be either linked at the beginning or at  
the end of a power supply serial link system.  
At the beginning of the serial link system, the mainframe is directly connected to the controller via the GPIB cable.  
The frame mode switch on the back of this mainframe is set to "MAIN". The mainframe has a unique primary GPIB  
address and the modules in this mainframe use secondary addresses (also referred to as subaddresses) 0 through 7. All  
power supplies serially linked to the mainframe use the same primary GPIB address, but must use secondary addresses  
8 through 15.  
At the end of the serial link system, the mainframe is linked to the last power supply in the system with the serial-link  
cable. The frame mode switch on the back of this mainframe is set to "AUX". The mainframe uses the same primary  
GPIB address as the power supplies and the modules in this mainframe use secondary addresses 8 through 15. In this  
configuration, you can have up to eight power supplies at secondary addresses 0 through 7 ahead of the mainframe.  
Keyboard Connection  
The optional Agilent 66001A MPS Keyboard connects to the mainframe through the keyboard jack located at the front of  
the mainframe and at the back of the mainframe as shown in Figure 3-1. More information on using the keyboard can be  
found in the module User’s Guide.  
Digital Connections  
A 4-pin connector and a quick-disconnect mating plug provide a discrete fault indicator (FLT) output and a remote inhibit  
(INH) input (see Figure 3-3). Disconnect the mating plug to make your wire connections. The electrical characteristics of  
the FLT output and the INH input are described in Chapter 1. More information on using digital connectors can be found in  
Chapter 4 of the module User’s Guideunder "RI Function" and "DFI Function".  
22  
Connections  
Figure 3-2. Controller Connections  
23  
Connections  
Figure 3-3. Digital Connector  
FLT Output  
(pins 1 and 2)  
Used to indicate that a fault has occurred on one of the modules in the mainframe. Pins 1 and 2  
are the open-collector output of an optocoupler, with pin 1 the collector and pin 2 the emitter. When a  
fault has occurred, pin 1 is driven low with respect to pin 2 (negative-true).  
INH Input  
(pin 3)  
Used to shut down the outputs of all modules in the mainframe that have been enabled to respond to  
the remote inhibit function. Pin 3 is a high impedance input. The inhibit function is activated when this  
input is driven low (negative-true). This can be accomplished by shorting pin 3 to pin 4.  
INH Common  
Provides the common connection for the INH input.  
(pin 4)  
After you have finished making all wire connections, replace the mating plug in the connector.  
Figure 3-4 shows two simple applications for the INH input. The first is a "panic switch" that applies the INH input signal  
to both mainframes, which shuts down the modules simultaneously in both mainframes. The second application uses the  
INH input signal to shut down modules in the main mainframe. In this case, the modules in the main mainframe must be  
programmed to generate an FLT signal upon receipt of the INH signal. The FLT signal generated by the modules in the  
main mainframe is then applied to the RI input of the auxiliary mainframe.  
Note  
Software control of the RI function is described further in Chapter 5 of the module Programming Guide.  
Trigger Connections  
Trigger connections are made to the BNC connectors on the rear panel as shown in Figure 3-1. The electrical  
characteristics of the trigger connectors are described in Chapter 1. More information on the capabilities of the external  
triggers can be found in Chapter 5 of the module Programming Guide.  
Trigger IN  
Allows negative-going external trigger signals to trigger one or more modules in the mainframe.  
Trigger OUT  
Provides a negative-going wired-ORed pulse that indicates a trigger has occurred on one or more of the  
modules in the mainframe.  
24  
Connections  
Figure 3-4. Examples of the Remote Inhibit Function  
25  
Connections  
Index  
A
M
addresses, 16  
MAIN, 16, 22  
address switch, 16  
manuals, 9, 13  
Agilent 603xA, 10, 11, 16, 22  
Agilent 664xA, 10, 11, 16, 22  
Agilent 665xA, 10, 11, 16, 22  
Agilent 667xA, 10, 11, 16, 22  
AUX, 16, 17, 22  
module, 11, 21, 22  
module - addresses, 22  
module - connector, 21  
module - installation, 18  
module - removal, 18-19  
modules, 11, 18  
B
BNC connectors, 24  
O
options, 10  
C
characteristics, 11  
connector, 18, 19, 21  
connector - installation, 21  
P
parts, 10  
power cord, 10, 13, 14  
power cord – installation, 14  
power plug, 10, 14  
primary address, 22  
pull tab, 18, 19  
D
digital connector plug, 10, 22, 24  
documentation, 9  
F
R
feet, 15  
rack mounting, 15  
filler panel, 10, 13, 18  
FLT/INH, 12, 24  
frame mode, 17, 22  
frame switch, 16, 22  
fuses, 14  
S
safety cover, 14  
screws, 10  
secondary address, 16, 22  
serial cable, 10, 13, 22  
serial-link, 22  
H
GPIB address, 16, 22  
GPIB interface, 11  
serial number, 18  
slot address, 16, 17, 18  
specifications, 11  
strain relief, 14, 15  
support rails, 15  
switches, 16  
K
keyboard, 22  
L
line current, 11  
line voltage, 11, 13  
line voltage, changing, 13  
T
temperature range, 15  
terminal block, 14  
trigger - connections, 24  
Trigln/TrigOut, 11, 24  
W
warranty, 13  
27  
Index  
Agilent Sales and Support Offices  
For more information about Agilent Technologies test and measurement products, applications, services,  
You can also contact one of the following centers and ask for a test and measurement sales  
representative.  
United States:  
Latin America:  
Agilent Technologies  
Test and Measurement Call Center  
P.O. Box 4026  
Englewood, CO 80155-4026  
(tel) 1 800 452 4844  
Agilent Technologies  
Latin American Region Headquarters  
5200 Blue Lagoon Drive, Suite #950  
Miami, Florida 33126  
U.S.A.  
(tel) (305) 267 4245  
(fax) (305) 267 4286  
Canada:  
Australia/New Zealand:  
Agilent Technologies Canada Inc.  
5150 Spectrum Way  
Mississauga, Ontario  
L4W 5G1  
Agilent Technologies Australia Pty Ltd  
347 Burwood Highway  
Forest Hill, Victoria 3131  
(tel) 1-800 629 485 (Australia)  
(fax) (61 3) 9272 0749  
(tel) 1 877 894 4414  
(tel) 0 800 738 378 (New Zealand)  
(fax) (64 4) 802 6881  
Europe:  
Asia Pacific:  
Agilent Technologies  
Test & Measurement European Marketing Organisation  
P.O. Box 999  
1180 AZ Amstelveen  
The Netherlands  
Agilent Technologies  
24/F, Cityplaza One, 1111 King’s Road,  
Taikoo Shing, Hong Kong  
tel: (852)-3197-7777  
fax: (852)-2506-9284  
(tel) (31 20) 547 9999  
Japan:  
Agilent Technologies Japan Ltd.  
Measurement Assistance Center  
9-1, Takakura-Cho, Hachioji-Shi,  
Tokyo 192-8510, Japan  
(tel) (81) 426 56 7832  
(fax) (81) 426 56 7840  
Technical data is subject to change.  
Manual Updates  
The following updates have been made to this manual since the print revision indicated on the title page.  
4/15/00  
All references to HP have been changed to Agilent.  
All references to HP-IB have been changed to GPIB.  

HP Hewlett Packard Hewlett Packard Computer Monitor 2011S User Manual
Hotpoint Cooktop CH 10450 GF S User Manual
GE DECT 28112xx2 User Manual
GE DECT 6 0 CORDLESS HANDSET 28111 User Manual
Fujitsu TE988G User Manual
Electrolux Air O Steam AOS201EAB1 User Manual
Electrolux 260510 User Manual
Dell Computer Monitor IN2020M User Manual
DCS RGS 366 User Manual
Bravetti Convection Oven CO200B User Manual