Teledyne Slow Cooker Model 2010B User Manual

OPERATING INSTRUCTIONS FOR  
Model 2010B  
Thermal Conductivity Analyzer  
Teledyne Analytical Instruments  
0.0  
AL-1  
%
Anlz  
2010B  
Thermal Conductivity Analyzer  
DANGER  
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING  
SYSTEM.  
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.  
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST  
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.  
P/N M70845  
11/22/00  
ECO # 00-0517  
ONLYAUTHORIZEDPERSONNELSHOULDCONDUCTMAINTENANCEAND/ORSERVICING. BEFORE  
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/  
MANAGER.  
Teledyne Analytical Instruments  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Table of Contents  
Specific Model Information ................................. iv  
Part I: Control Unit, Model 2010B .........Part I: 1-1  
Part II: Analysis Unit, Model 2010B......Part II: 1-1  
Appendix ......................................................... A-1  
Teledyne Analytical Instruments  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 2010B  
Specific Model Information  
The instrument for which this manual was supplied may incorporate one or more  
options not supplied in the standard instrument. Commonly available options are listed  
below, with check boxes. Any that are incorporated in the instrument for which this  
manual is supplied are indicated by a check mark in the box.  
Instrument Serial Number: _______________________  
Options Included in the Instrument with the Above Serial Number:  
C:  
Auto Calibration valves (zero/span/sample) built-in gas control  
valves are electronically controlled to provide synchronization  
with the analyzer’s operations.  
G:  
H:  
K:  
Stainless steel cell block with nickel filaments and Stainless  
Steel fittings and tubing.  
Stainless steel cell block with gold filaments for corrosive gas  
streams and Stainless Steel fittings and tubing.  
19" Rack Mount option with one or two analyzer Control  
Units installed and ready to mount in a standard rack.  
K2:  
19" Rack Mount option with two Control Units mounted.  
K3:  
19" Rack Mount option with one Control Unit mounted and a  
blank cover installed in the second Control Unit location.  
L:  
Gas selector panel consisting of sample/ref flow meters and  
control valves for metering input of sample/calibrations support  
gases.  
F:  
Flame Arrestors for Class 1, Div. 1, Groups C/D service.  
P:  
Flame Arrestors for Class 1, Div. 1, Groups C/D service, and  
Auto Cal valves option (Ref. C above) and GP use.  
Q:  
Flame Arrestors for Group B (hydrogen) service, and Auto Cal  
valves option (Ref. C above)  
O:  
Flame Arrestors for Group B (hydrogen)  
Sealed Reference Cell  
R:  
iv  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Table of Contents  
1 Introduction  
1.1 Overview ........................................................................ 1-1  
1.2 Typical Applications ....................................................... 1-2  
1.3 Main Features of the Analyzer ....................................... 1-2  
1.4 Model Designations ....................................................... 1-3  
1.5 Operator Interface (Front Panel) .................................... 1-3  
1.6 Recognizing Difference Between LCD & VFD ............... 1-5  
1.7 Equipment Interface (Rear Panel).................................. 1-5  
1.8 Gas Connections ........................................................... 1-7  
2 Operational Theory  
2.1 Introduction .................................................................... 2-1  
2.2 Sensor Theory ............................................................... 2-1  
2.2.1 Sensor Configuration .............................................. 2-1  
2.2.2 Calibration ............................................................... 2-2  
2.2.3 Effects of Flowrate and Gas Density ....................... 2-3  
2.2.4 Measurement Results ............................................. 2-3  
2.3 Electronics and Signal Processing ................................ 2-3  
2.4 Temperature Control ...................................................... 2-5  
3 Installation  
3.1 Unpacking the Analyzer ................................................. 3-1  
3.2 Mounting the Control Unit .............................................. 3-1  
3.3 Electrical Connections (Rear Panel) .............................. 3-3  
3.3.1 Primary Input Power .............................................. 3-4  
3.3.2 Fuse Installation..................................................... 3-4  
3.3.3 Analog Outputs ...................................................... 3-4  
3.3.4 Alarm Relays ......................................................... 3-6  
3.3.5 Digital Remote Cal Inputs ...................................... 3-7  
3.3.6 Range ID Relays.................................................... 3-8  
3.3.7 Network I/O............................................................ 3-8  
3.3.8 RS-232 Port ........................................................... 3-9  
3.3.9 Remote Probe Connector ...................................... 3-9  
3.4 Testing the System ........................................................3-16  
3.5 Warm Up at Power Up ...................................................3-16  
4 Operation  
4.1 Introduction .................................................................... 4-1  
4.2 Using the Data Entry and Function Buttons................... 4-1  
Teledyne Analytical Instruments  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 2010B  
4.3 The System Function ..................................................... 4-4  
4.3.1 Setting the Display ................................................. 4-5  
4.3.2 Setting up an Auto-Cal........................................... 4-5  
4.3.3 Password Protection .............................................. 4-6  
4.3.3.1 Entering the Password................................... 4-7  
4.3.3.2 Installing or Changing the Password ............. 4-7  
4.3.4 Logging Out ........................................................... 4-9  
4.3.5 System Self-Diagnostic Test .................................. 4-9  
4.3.6 The Model Screen ................................................. 4-10  
4.3.7 Checking Linearity with Algorithm.......................... 4-10  
4.4 The Zero and Span Functions ....................................... 4-11  
4.4.1 Zero Cal ................................................................. 4-12  
4.4.1.1 Auto Mode Zeroing ........................................ 4-12  
4.4.1.2 Manual Mode Zeroing.................................... 4-13  
4.4.1.3 Cell Failure..................................................... 4-14  
4.4.2 Span Cal ................................................................ 4-14  
4.4.2.1 Auto Mode Spanning ..................................... 4-15  
4.4.2.2 Manual Mode Spanning................................. 4-15  
4.5 The Alarms Function...................................................... 4-16  
4.6 The Range Select Function ........................................... 4-18  
4.6.1 Manual (Select/Define Range) Screen .................. 4-19  
4.6.2 Auto (Single Application) Screen ........................... 4-19  
4.6.3 Precautions............................................................ 4-21  
4.7 The Analyze Function .................................................... 4-21  
4.8 Programming ................................................................. 4-22  
4.8.1 The Set Range Screen .......................................... 4-23  
4.8.2 The Curve Algorithm Screen ................................. 4-25  
4.8.2.1 Checking the Linearization ............................ 4-25  
4.8.2.2 Manual Mode Linearization............................ 4-26  
4.8.2.3 Auto Mode Linearization ................................ 4-27  
4.9 Special Function Setup .................................................. 4-28  
4.9.1 Output Signal Reversal .......................................... 4.28  
4.9.2 Special - Inverting Output ...................................... 4-28  
4.9.3 Special - Polarity Coding........................................ 4-29  
4.9.4 Special - Nonlinear Application Gain Preset .......... 4-29  
Maintenance  
5.1 Routine Maintenance ..................................................... 5-1  
5.2 System Self Diagnostic Test........................................... 5-1  
5.3 VFD Display ................................................................... 5-2  
5.4 Fuse Replacement......................................................... 5-2  
vi  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
5.5 Major Internal Components ........................................... 5-3  
5.6 Cleaning......................................................................... 5-5  
5.7 Phone Numbers ............................................................. 5-5  
Appendix  
A-1 Specifications................................................................. A-1  
A-2 Recommended 2-Year Spare Parts List......................... A-3  
A-3 Drawing List ................................................................... A-4  
Teledyne Analytical Instruments  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 2010B  
DANGER  
COMBUSTIBLE GAS USAGE WARNING  
The customer should ensure that the principles of operating of  
this equipment are well understood by the user. Misuse of this  
product in any manner, tampering with its components, or  
unauthorized substitution of any component may adversely  
affect the safety of this instrument.  
Since the use of this instrument is beyond the control of  
Teledyne, no responsibility by Teledyne, its affiliates, and  
agents for damage or injury from misuse or neglect of this  
equipment is implied or assumed.  
viii  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Introduction  
1.1 Overview  
TheModel2010isafamilyofsplitconfigurationconductivityanalyzers.  
EachanalyzerconsistofaControlUnitsuitableforinstallationinageneral  
purposearea-enclosureratedNEMA-4,andaAnalysisUnitwhichishousedin  
anexplosion-proofenclosure. TheAnalysisUnitenclosureisratedforNEMA  
4/7 Class I, Div. 1, Groups B,C,D and is approved by U/L and CSA.  
TheAnalyticalInstrumentsModel2010BThermalConductivityAnalyzeris  
aversatilemicroprocessor-basedinstrumentformeasuringacomponentgasina  
backgroundgas,orinaspecificmixtureofbackgroundgases.Itcomparesthe  
thermalconductivityofasamplestreamwiththatofareferencegasofknown  
composition.The2010Bcan—  
measure the concentration of one gas in a mixture of two gases.  
measure the concentration of a gas in a specific mixture of back-  
ground gases.  
measure the purity of a sample stream containing a single impuri-  
ty or a mixture of impurities.  
Thestandard2010Bispre-programmedwithautomaticlinearization  
algorithmsforalargenumberofgasesandgasmixtures.Thefactorycanaddto  
thisdatabaseforcustomapplications,andthesophisticatedusercanaddhis  
ownuniqueapplications.  
ThismanualsectioncoverstheModel2010BGeneralPurposeflush-panel  
andrack-mountcontrolunitsonly.  
ManyoftheModel2010Bfeaturescoveredinthismanualareoptional,  
selectedaccordingtothecustomersspecificapplication.Refertothespecific  
modelinformationsheet(pageIV)fortheoptionsincorporatedintheinstrument.  
Teledyne Analytical Instruments  
Part I 1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 2010B  
1.2 Typical Applications  
AfewtypicalapplicationsoftheModel2010Bare:  
Power generation  
Airliquefaction  
Chemicalreactionmonitoring  
Steel manufacturing and heat treating  
Petrochemicalprocesscontrol  
Qualityassurance  
Refrigerationandstorage  
Gasproportioningcontrol.  
1.3 Main Features of the Analyzer  
ThemainfeaturesoftheModel2010BThermalConductivityAnalyzer  
include:  
Three independent, user definable, analysis ranges allow up to  
three different gas applications with one concentration range  
each, or up to three concentration ranges for a single gas applica-  
tion, or any combination.  
Specialrecalibrationrangeformultipleapplications.Recalibrat-  
ing one, recalibrates all.  
Automatic, independent linearization for each range.  
Auto Ranging allows analyzer to automatically select the proper  
preset range for a given single application. Manual override  
allows the user to lock onto a specific range of interest.  
RS-232 serial digital port for use with a computer or other digital  
communicationsdevice.  
Six adjustable concentration set points with two alarms and a system  
failurealarmrelay.  
Extensive self-diagnostic testing, at startup and on demand.  
A 2-line alphanumeric display screen, driven by microprocessor  
electronics, that continuously prompts and informs the operator.  
High resolution, accurate indication of target or impurity gas  
concentration from large, bright, meter readout. (0-9999 ppm  
through 0-100 % depending on types of gas involved.)  
Standard, proven sensor cell design.  
1-2 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Wide range of custom applications, ranges, and linearization.  
Microprocessorbasedelectronics:8-bitCMOSmicroprocessor  
with 32 kB RAM and 128 kB ROM.  
Autoandremotecalibrationcapabilities.  
Four analog outputs: two for measurement (0–1 V dc and Isolat-  
ed 4–20 mA dc) and two for range identification.  
Compact and versatile design: Small footprint, yet internal com-  
ponentsareaccessible.  
1.4 Model Designations  
TheModel2010Bisordinarilycustomprogrammedatthefactorytofitthe  
customer’sapplication.Manyparameters,includingthenumberofchannels,the  
gasapplication,thematerialsspecificationofthesamplingsystem,andothers,  
areoptions.Themostcommonoptions,arecoveredinthismanual.Seethe  
SpecificModelInformationchecklistinthefrontmatterofthismanualforthose  
thatapplytoyourModel2010Banalyzer.Somestandardmodelsthatarenot  
coveredinthismanualarelistedhere.  
Models 2000A: Both analysis section and control unit are in a single  
generalpurposeenclosure.  
Models 2020:  
Both the analysis section and control unit are in a single  
explosionproofenclosure.  
1.5 Operator Interface (Front Panel)  
Thestandard2010Bishousedinaruggedmetalcasewithallcontrolsand  
displaysaccessiblefromthefrontpanel.SeeFigure1-1.Thefrontpanelhas  
thirteenbuttonsforoperatingtheanalyzer,adigitalmeter,andanalphanumeric  
display.TheyaredescribedbrieflyhereandindetailintheOperationschapter  
ofthismanual.  
FunctionKeys:Sixtouch-sensitivemembraneswitchesareusedto  
changethespecificfunctionperformedbytheanalyzer:  
Analyze Perform analysis for target-gas content of a sample  
gas.  
System Performsystem-relatedtasks(describedindetailin  
chapter 4, Operation.).  
Teledyne Analytical Instruments  
Part I 1-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 2010B  
Outer Door  
(Open)  
Teledyne Analytical Instruments  
Digital  
M eter  
Viewing  
W indow  
0.0  
AL-1  
%
Anlz  
LCD  
Screen  
Inner Door  
Latch  
(Pressing the latch button  
will open the inner Door)  
Outer Door  
Latch  
Control  
Panel  
2010B  
Therm al Conductivity Analyzer  
Figure 1-1: Model 2010B Front Panel  
Span  
Zero  
Alarms Set the alarm setpoints and attributes.  
Range Set up the user definable ranges for the instrument.  
Span calibrate the analyzer.  
Zero calibrate the analyzer.  
DataEntryKeys:Sixtouch-sensitivemembraneswitchesareusedto  
inputdatatotheinstrumentviathealphanumericVFD(VacuumFluorescent  
Display)display:  
Left & Right Arrows  
Selectbetweenfunctionscurrently  
displayed on the VFD screen.  
Up & Down Arrows  
Increment or decrement values of  
functionscurrentlydisplayed.  
Enter Moves VFD on to the next screen in a series. If none  
remains, returns to the Analyze screen.  
Escape Moves VFD back to the previous screen in a series. If  
none remains, returns to the Analyze screen.  
1-4 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Digital Meter Display: ThemeterdisplayisaVFDdevicethat  
produceslarge,bright,7-segmentnumbersthatarelegibleinanylighting.It  
producesacontinuoustracereadoutfrom0-9999ppmoracontinuouspercent  
readoutfrom1-100 %.Itisaccurateacrossallanalysisranges.  
AlphanumericInterfaceScreen:TheVDFscreenisaneasy-to-use  
interfacebetweenoperatorandanalyzer.Itdisplaysvalues,options,and  
messagesthatgivetheoperatorimmediatefeedback.  
StandbyButton:The Standbyturnsoffthedisplayandoutputs,  
butcircuitryisstilloperating.  
CAUTION: The power cable must be unplugged to fully  
disconnect power from the instrument. When  
chassis is exposed or when access door is open  
and power cable is connected, use extra care to  
avoid contact with live electrical circuits.  
AccessDoor:Foraccesstothethermalconductivitysensororthefront  
panelelectronics,thefrontpanelswingsopenwhenthelatchintheupperright  
cornerofthepanelispressedallthewayinwithanarrowgaugetool.Accessing  
themainelectronicscircuitboardrequiresunfasteningrearpanelscrewsand  
slidingtheelectronicsdraweroutofthecase.(Seechapter5.)  
1.6 Recognizing Difference Between LCD &  
VFD  
LCD(LiquidCrystalDisplay)hasGREENbackgroundwithBLACK  
characters. VFD has DARK background withGREEN characters. In the case  
ofVFD (VacuumFluorescentDisplay)-NOCONTRASTADJUSTMENTIS  
NEEDED.  
1.7 Equipment Interface (Rear Panel)  
Therearpanel,showninFigure1-2,containstheelectricalconnectorsfor  
externalinputandoutput.Theconnectorsaredescribedbrieflyhereandindetail  
inchapter3,Installation.  
Teledyne Analytical Instruments  
Part I 1-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 2010B  
Figure 1-2: Model 2010B Rear Panel  
Power Connection  
Analog Outputs  
85-250 V AC power source.  
0-1 V dc concentration plus 0-1 V dc  
range ID, and isolated 4-20 mA dc plus  
4-20 mA dc range ID.  
1-6 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Alarm Connections  
2 concentration alarms and 1 system  
alarm.  
RS-232 Port  
Serialdigitalconcentrationsignaloutput  
and control input.  
Remote Probe  
Remote Span/Zero  
Used in the 2010B to interface the  
external Analysis Unit.  
Digital inputs allow external control of  
analyzercalibration.  
Calibration Contact To notify external equipment that  
instrument is being calibrated and  
readings are not monitoring sample.  
Range ID Contacts  
Network I/O  
Four separate, dedicated,  
range-identificationrelaycontacts(01,  
02, 03,CAL).  
Serialdigitalcommunicationsforlocal  
network access. For future expansion.  
Not implemented at this printing.  
Note: If you require highly accurate Auto-Cal timing, use external  
Auto-Cal control where possible. The internal clock in the  
Model 2010B is accurate to 2-3 %. Accordingly, internally  
scheduled calibrations can vary 2-3 % per day.  
Teledyne Analytical Instruments  
Part I 1-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 2010B  
1-8 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Operational Theory  
2.1 Introduction  
The analyzer is composed of two subsystems:  
1. ThermalConductivitySensor  
2. Electronic Signal Processing, Display and Control.  
The sensor is a thermal conductivity comparator that continuously  
compares the thermal conductivity of the sample gas with that of a reference  
gas having a known conductivity.  
The electronic signal processing, display and control subsystem simpli-  
fies operation of the analyzer and accurately processes the sampled data. A  
microprocessorcontrolsallsignalprocessing, input/output, anddisplay  
functions for the analyzer.  
2.2 Sensor Theory  
For greater clarity, Figure 2-1 presents two different illustrations, (a)  
and (b), of the operating principle of the thermal conductivity cell.  
2.2.1 Sensor Configuration  
The thermal conductivity sensor contains two chambers, one for the  
reference gas of known conductivity and one for the sample gas. Each  
chamber contains a pair of heated filaments. Depending on its thermal  
conductivity, each of the gases conducts a quantity of heat away from the  
filaments in its chamber. See Figure 2-1(a).  
The resistance of the filaments depends on their temperature. These  
filaments are parts of the two legs of a Wheatstone bridge circuit that unbal-  
ances if the resistances of its two legs do not match. See Figure 2-1(b).  
Teledyne Analytical Instruments  
Part I 2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2 Operational Theory  
Model 2010B  
Figure 2-1: Thermal Conductivity Cell Operating Principle  
If the thermal conductivities of the gases in the two chambers are  
different, the Wheatstone bridge circuit unbalances, causing a current to flow  
in its detector circuit. The amount of this current can be an indication of the  
amount of impurity in the sample gas, or even an indication of the type of  
gas, depending on the known properties of the reference and sample gases.  
The temperature of the measuring cell is regulated to within 0.1 °C by a  
sophisticated control circuit. Temperature control is precise enough to com-  
pensate for diurnal effects in the output over the operating ranges of the  
analyzer. (See Specifications in the Appendix for details.)  
2.2.2 Calibration  
Because analysis by thermal conductivity is not an absolute measure-  
ment, calibration gases of known composition are required to fix the upper  
and lower parameters (“zero” and “span”) of the range, or ranges, of analy-  
sis. These gases must be used periodically, to check the accuracy of the  
analyzer.  
During calibration, the bridge circuit is balanced, with zero gas against  
the reference gas, at one end of the measurement range; and it is sensitized  
with span gas against the reference gas at the other end of the measurement  
range. The resulting electrical signals are processed by the analyzer electron-  
ics to produce a standard 0-1V, or an isolated 4–20 mA dc, output signal, as  
described in the next section.  
Teledyne Analytical Instruments  
2-2 Part I  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
2.2.3 Effects of Flowrate and Gas Density  
Because the flowrate of the gases in the chambers affects their cooling  
of the heated filaments, the flowrate in the chambers must be kept as equal,  
constant, and low as possible.  
When setting the sample and reference flowrate, note that gases lighter  
than air will have an actual flowrate higher than indicated on the flowmeter,  
while gases heavier than air will have an actual flowrate lower than indi-  
cated. Due to the wide range of gases that are measured with the Thermal  
Conductivity Analyzer, the densities of the gases being handled may vary  
considerably.  
Then, there are limited applications where the reference gas is in a  
sealed chamber and does not flow at all. These effects must be taken in  
consideration by the user when setting up an analysis.  
2.2.4 Measurement Results  
Thermal conductivity measurements are nonspecific by nature. This fact  
imposes certain limitations and requirements. If the user intends to employ  
the analyzer to detect a specific component in a sample stream, the sample  
must be composed of the component of interest and one other gas (or spe-  
cific, and constant, mixture of gases) in order for the measured heat-transfer  
differences to be nonambiguous.  
If, on the other hand, the user is primarily interested in the purity of a  
process stream, and does not require specific identification of the impurity,  
the analyzer can be used on more complex mixtures.  
2.3 Electronics and Signal Processing  
The Model 2010B Thermal Conductivity Analyzer uses an 8031  
microcontroller, Central Processing Unit—(CPU) with 32 kB of RAM and  
128 kB of ROM to control all signal processing, input/output, and display  
functions for the analyzer. System power is supplied from a universal power  
supply module designed to be compatible with any international power  
source. (See Major Internal Components in chapter 5 Maintenance for the  
location of the power supply and the main electronic PC boards.)  
The signal processing electronics including the microprocessor, analog  
to digital, and digital to analog converters are located on the Motherboard at  
the bottom of the case. The Preamplifier board is mounted on top of the  
Motherboard as shown in the figure 5.4. These boards are accessible after  
Teledyne Analytical Instruments  
Part I 2-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2 Operational Theory  
Model 2010B  
removing the back panel. Figure 2-2 is a block diagram of the Analyzer  
electronics.  
Variable  
Differential  
Amplifier  
Gain  
Amplifier  
Thermistor  
A to D  
Converter  
To CPU  
M
U
X
Sensor  
Heater  
Auto-  
Range  
0-1 V dc  
Concentration  
and Range  
Temperature  
Control  
Digitial to  
Analog  
Converter  
(DAC)  
Coarse  
Adjustment  
4-20 mA dc  
Concentration  
and Range  
Fine  
Adjustment  
Temperature  
Control  
Heater  
Alarm 1  
Alarm 2  
Selt-Test Signal to MUX  
System  
Failure  
Alarm  
RS-232  
Keyboard  
Central  
Processing  
Unit  
Range  
Contacts (4)  
(CPU)  
Displays  
Processing  
External  
Valve  
Control  
Remote Span  
Control  
Power  
Supply  
Remote Zero  
Control  
A to D Conv  
Cal  
Contact  
Figure 2-2: Block Diagram of the Model 2010B Electronics  
Teledyne Analytical Instruments  
2-4 Part I  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
In the presence of dissimilar gases the sensor generates a differential  
voltageacrossitsoutputterminals. Adifferentialamplifierconvertsthis  
signal to a unipolar signal, which is amplified in the second stage, variable  
gain amplifier, which provides automatic range switching under control of  
the CPU. The output from the variable gain amplifier is sent to an 18 bit  
analog to digital converter.  
The digital concentration signal along with input from the Gas Selector  
Panel is processed by the CPU and passed on to the 12-bit DAC, which  
outputs 0-1 V dc Concentration and Range ID signals. An voltage-to-current  
converter provides 4-20 mA dc concentration signal and range ID outputs.  
The CPU also provides appropriate control signals to the Displays,  
Alarms, and External Valve Controls, and accepts digital inputs for external  
Remote Zero and Remote Span commands. It monitors the power supply  
through an analog to digital converter as part of the data for the system  
failurealarm.  
The RS-232 port provides two-way serial digital communications to  
and from the CPU. These, and all of the above electrical interface signals are  
described in detail in chapter 3 Installation.  
2.4. Temperature Control  
For accurate analysis the sensor of this instrument is temperature con-  
trolled to 50oC.  
The Temperature Control keeps the temperature of the measuring cell  
regulated to within 0.1 degree C. A thermistor is used to measure the tem-  
perature, and a zero-crossing switch regulates the power in a cartridge-type  
heater. The result is a sensor output signal that is temperature independent.  
A second temperature control system is used to maintain the Analysis  
Unit internal temperature at 220C minimum.  
Teledyne Analytical Instruments  
Part I 2-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2 Operational Theory  
Model 2010B  
Teledyne Analytical Instruments  
2-6 Part I  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Installation  
Installation of the Model 2010B Analyzer includes:  
1. Unpacking  
2. Mounting  
3. Gas connections  
4. Electricalconnections  
5. Testing the system.  
3.1 Unpacking the Analyzer  
The analyzer is shipped ready to install and prepared for operation.  
Carefully unpack the analyzer and inspect it for damage. Immediately report  
any damage to the shipping agent.  
3.2 Mounting the Control Unit  
The Model 2010B Control Unit is designed for bulkhead mounting in  
general purpose area, NOT for hazardous environments of any type. The  
Control Unit is for indoor/outdoor use.  
Figure 3-1 is an illustration of the 2010B standard front panel and  
mounting bezel. There are four mounting holes—one in each corner of the  
rigid frame. See the outline drawing, at the back of this manual for overall  
dimensions.  
All operator controls are mounted on the inner control panel, which is  
hinged on the left edge and doubles as a door to provide access to the internal  
components of the instrument. The door will swing open when the button of  
the latch is pressed all the way in with a narrow gauge tool (less than 0.18  
inch wide), such as a small hex wrench or screwdriver Allow clearance for  
the door to open in a 90-degree arc of radius 11.75 inches. See Figure 3-2.  
Teledyne Analytical Instruments  
Part I 3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 2010B  
NPT Fittings  
supplied by  
customer  
Viewing  
W indow  
0.0  
AL-1  
% Anlz  
Outer Door  
Latch  
HHinge  
Figure 3-1: Front Panel of the Model 2010B Control Unit  
Figure 3-2: Required Front Door Clearance  
H
3-2 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
3.3 Electrical Connections (Rear Panel)  
Figure 3-3 shows the Model 2010B Electrical Connector Panel. There  
are terminal blocks for connecting power, communications, and both digital  
and analog concentration outputs.  
For safe connections, ensure that no uninsulated wire extends outside of  
the connectors they are attached to. Stripped wire ends must insert com-  
pletely into terminal blocks. No uninsulated wiring should be able to come in  
contact with fingers, tools or clothing during normal operation.  
Figure 3-3: Rear Panel of the Model 2010B  
3.3.1 Primary Input Power  
The power terminal block and fuse receptacles are located in the same  
rearpanelassembly.  
Teledyne Analytical Instruments  
Part I 3-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 2010B  
DANGER: POWER IS APPLIED TO THE INSTRUMENT'S CIR-  
CUITRY AS LONG AS THE INSTRUMENT IS CON-  
NECTED TO THE POWER SOURCE. THE RED I/O  
SWITCH ON THE FRONT PANEL IS FOR SWITCH-  
ING POWER ON OR OFF TO THE DISPLAYS AND  
OUTPUTS ONLY.  
NOTE:  
AC POWER MAY BE PRESENT ON THE RELAY  
CONTACTS WHEN THE POWER CORD IS RE-  
MOVED!  
The Control Unit is universal power 100-240V, 50-60 Hz. The Analy-  
sis Unit requires 110 or 220 VAC and is selectable via switch located inside  
theexplosion-proofenclosure.  
3.3.2 Fuse Installation  
The fuse receptacles accepts European size fuses only (5x20mm). Both  
sides of the line are fused. If the European size fuses are selected, both sides  
at the line will be fused. Be sure to install the proper fuse as part of installa-  
tion. (See Fuse Replacement in chapter 5, maintenance.)  
3.3.3 Analog Outputs  
There are four DC output signal connectors with spring terminals on the  
panel. There are two wires per output with the polarity noted. See Figure 3-  
4. The outputs are:  
0–1 V dc % of Range: Voltageriseslinearlywithincreasingconcentration,  
from 0 V at 0 concentration to 1 V at full scale.  
(Full scale = 100% of programmable range.)  
0–1 V dc Range ID:  
0.25 V = Range 1, 0.5 V = Range 2, 0.75 V =  
Range 3, 1 V = Cal Range.  
4–20 mA dc % Range: Current rises linearly with concentration, from 4  
mA at 0 concentration to 20 mA at full scale. (Full  
scale = 100% of programmable range.)  
4–20 mA dc Range ID: 8 mA = Range 1, 12 mA = Range 2, 16 mA =  
Range 3, 20 mA = Range 4.  
3-4 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Figure 3-4: Analog Output Connections  
Examples:  
The analog output signal has a voltage which depends on gas concen-  
tration relative to the full scale of the range. To relate the signal output to the  
actual concentration, it is necessary to know what range the instrument is  
currently on, especially when the analyzer is in the autoranging mode.  
The signal output for concentration is linear over the currently selected  
analysis range. For example, if the analyzer is set on a range that was defined  
as 0–10 % hydrogen, then the output would be as shown in Table 3-1.  
Table 3-1: Analog Concentration Output—Example  
Percent  
Voltage Signal Current Signal  
Hydrogen Output (V dc)  
Output (mA dc)  
0
1
2
3
4
5
6
7
8
0.0  
0.1  
0.2  
0.3  
0.4  
0.5  
0.6  
0.7  
0.8  
0.9  
1.0  
4.0  
5.6  
7.2  
8.8  
10.4  
12.0  
13.6  
15.2  
16.8  
18.4  
20.0  
9
10  
(Continued)  
Teledyne Analytical Instruments  
Part I 3-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 2010B  
To provide an indication of the range, the Range ID analog output  
terminals are used. They generate a steady preset voltage (or current when  
using the current outputs) to represent a particular range. Table 3-2 gives the  
range ID output for each analysis range.  
Table 3-2: Analog Range ID Output—Example  
Range  
Voltage (V)  
Current (mA) Application  
Range 1  
0.25  
8
12  
16  
20  
0-1 % H2 in N2  
0-10 % H2 in N2  
0-1 % H2 in Air  
0-1 % H2 in N2  
Range 2  
0.50  
0.75  
1.00  
Range 3  
Range 4 (Cal)  
3.3.4 Alarm Relays  
The three alarm-circuit connectors are spring terminals for making  
connections to internal alarm relay contacts. Each provides a set of Form C  
contacts for each type of alarm. Each has both normally open and normally  
closed contact connections. The contact connections are indicated by dia-  
grams on the rear panel. They are capable of switching up to 3 amperes at  
250 V ac into a resistive load. See Figure 3-5. The connectors are:  
Threshold Alarm 1: • Can be configured as high (actuates when concen-  
tration is above threshold), or low (actuates when  
concentrationisbelowthreshold).  
• Can be configured as fail-safe or non-fail-safe.  
• Can be configured as latching or nonlatching.  
• Can be configured out (defeated).  
Threshold Alarm 2: • Can be configured as high (actuates when concen-  
tration is above threshold), or low (actuates when  
concentrationisbelowthreshold).  
• Can be configured as fail-safe or non-fail-safe.  
• Can be configured as latching or nonlatching.  
• Can be configured out (defeated).  
SystemAlarm:  
Actuates when DC power supplied to circuits is  
unacceptable in one or more parameters. Permanently  
configured as fail-safe and latching. Cannot be de-  
feated.  
Actuates when cell can not balance during zero  
calibration.  
Actuates when span parameter out off its limited  
parameter.  
3-6 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Actuates when self test fails.  
(Reset by pressing I/O button to remove power. Then  
press I/O again and any other button EXCEPT  
Systemtoresume.  
Further detail can be found in chapter 4, section 4-5.  
Figure 3-5: Types of Relay Contacts  
3.3.5 Digital Remote Cal Inputs  
Accept 0 V (off) or 24 V dc (on) inputs for remote control of calibra-  
tion. (See Remote Calibration Protocol below.)  
Zero:  
Floating input. 5 to 24 V input across the + and – terminals  
puts the analyzer into the Zero mode. Either side may be  
grounded at the source of the signal. A synchronous signal  
must open and close the external gas control valves appropri-  
ately. See 3.3.9 Remote Probe Connector. (With the –C  
option,theinternalvalvesoperateautomatically.)  
Span:  
Floating input. 5 to 24 V input across the + and – terminals  
puts the analyzer into the Span mode. Either side may be  
grounded at the source of the signal. A synchronous signal  
must open and close the external gas control valves appropri-  
ately. See 3.3.9 Remote Probe Connector. (With the –C  
option,theinternalvalvesoperateautomatically.)  
Cal Contact: This relay contact is closed while analyzer is spanning  
and/or zeroing. (See Remote Calibration Protocol below.)  
Remote Calibration Protocol: To properly time the Digital Remote  
Cal Inputs to the Model 2010B Analyzer, the customer's controller must  
monitor the Cal Relay Contact.  
Teledyne Analytical Instruments  
Part I 3-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 2010B  
When the contact is OPEN, the analyzer is analyzing, the Remote Cal  
Inputs are being polled, and a zero or span command can be sent.  
When the contact is CLOSED, the analyzer is already calibrating. It  
will ignore your request to calibrate, and it will not remember that request.  
Once a zero or span command is sent, and acknowledged (contact  
closes), release it. If the command is continued until after the zero or span is  
complete, the calibration will repeat and the Cal Relay Contact (CRC) will  
closeagain.  
When the contact is closed, the display would display the last reading of  
the gas concentration value and output signal would output the last reading  
from the sample gas (SAMPLE and HOLD).  
Forexample:  
1) Test the CRC. When the CRC is open, Send a zero command  
until the CRC closes (The CRC will close quickly.)  
2) When the CRC closes, remove the zero command.  
3) When CRC opens again, send a span command until the CRC  
closes. (The CRC will close quickly.)  
4) When the CRC closes, remove the span command.  
When CRC opens again, zero and span are done, and the sample is  
being analyzed.  
Note: The Remote Probe connector (paragraph 3.3.9) provides  
signals to operate the zero and span gas valves synchro-  
nously. However, if you have the –C, -P, or -Q Internal valve  
option, which includes zero and span gas inputs, the 2010B  
automatically selects the zero, span and sample gas flow.  
3.3.6 Range ID Relays  
Four dedicated Range ID relay contacts. For any single application they  
are assigned to relays in ascending order. For example: if all ranges have the  
same application, then the lowest range is assigned to the Range 1 ID relay,  
and the highest range is assigned to the Range 3 ID relay. Range 4 is the Cal  
Range ID relay.  
3.3.7 NetworkI/O  
A serial digital input/output for local network protocol. At this printing,  
this port is not yet functional. It is to be used in future versions of the instru-  
ment.  
3-8 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
3.3.8 RS-232 Port  
Part I: Control Unit  
The digital signal output is a standard RS-232 serial communications  
port used to connect the analyzer to a computer, terminal, or other digital  
device. It requires a standard 9-pin D connector.  
Output: The data output is status information, in digital form, updated  
every two seconds. Status is reported in the following order:  
The concentration in ppm or percent  
The range in use (01 = Range 1, 02 = Range 2, 03 = Range 3,  
CAL = Range 4)  
The scale of the range (0-100 %, etc)  
Which alarms—if any—are disabled (AL–x DISABLED)  
Which alarms—if any—are tripped (AL–x ON).  
Each status output is followed by a carriage return and line feed.  
Input: The input functions using RS-232 that have been implemented  
to date are described in Table 3-3.  
Table 3-3: Commands via RS-232 Input  
Command  
as<enter>  
az<enter>  
rp<enter>  
Description  
Immediatelystartsanautospan.  
Immediatelystartsanautozero.  
Allows reprogramming of the APPLICATION (gas use)  
and ALGORITHM (linearization) System functions.  
st<enter>  
Toggling input. Stops/Starts any status message output from  
the RS-232, until st<enter> is sent again.  
rm1<enter>  
rm2<enter>  
rm3<enter>  
rm4<enter>  
ra<enter>  
Range manual 1  
Range manual 2  
Range manual 3  
Range manual CAL  
Range auto  
Implementation: The RS-232 protocol allows some flexibility in its  
implementation. Table 3-4 lists certain RS-232 values that are required by  
the Model 2010B implementation.  
Teledyne Analytical Instruments  
Part I 3-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 2010B  
Table 3-4: Required RS-232 Options  
Parameter  
Setting  
Baud 2400  
Byte  
Parity  
Stop Bits  
8 bits  
none  
1
MessageInterval  
2 seconds  
3.3.9 Remote Probe Connector  
The 2010B is a single-split configuration analyzer, the Remote Probe  
connector is used to interface the Analysis Unit. See Figure 3-6.  
SPA N IN  
SAM PLE IN  
ZE RO IN  
EXH AUST  
+15V  
Turn cw to hold  
ccw to  
loosen w ire.  
Switching  
G round  
Insert wire  
here.  
Figure 3-6: Remote Probe Connector Pinouts  
The voltage from these outputs is nominally 0 V for the OFF and  
15 V dc for the ON conditions. The maximum combined current that can be  
pulled from these output lines is 100 mA. (If two lines are ON at the same  
time, each must be limited to 50 mA, etc.) If more current and/or a different  
voltage is required, use a relay, power amplifier, or other matching circuitry  
to provide the actual driving current.  
In addition, each individual line has a series FET with a nominal ON  
resistance of 5 ohms (9 ohms worst case). This could limit the obtainable  
voltage, depending on the load impedance applied. See Figure 3-7.  
3-10 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Figure 3-7: FET Series Resistance  
3.4 Testing the System  
Before plugging the instrument into the power source:  
Check the integrity and accuracy of the gas connections. Make  
sure there are no leaks.  
Check the integrity and accuracy of the electrical connections.  
Make sure there are no exposed conductors  
Check that the pressure and flow of all gases are within the  
recommended levels, and appropriate for your application.  
Power up the system, and test it by performing the following  
operations:  
1. Repeat the Self-Diagnostic Test as described in chapter 4, section  
4.3.5.  
3.5 Warm Up at Power Up  
Every time the unit is turned on, the instrument stays with the introduc-  
tion screen for thirty minutes. This is to allow the cell to come up to tem-  
perature (50oC). The only way to bypass this warm up period is by pressing  
any key once, such as the Enter key.  
The instrument warms up for half an hour so that it will not receive a  
remote calibration signal, send false readings to a monitor system, or, again,  
be calibrated by an untrained operator while the cell is cold.  
NOTE:There is no feedback on whether the working temperature  
has been achieved by cell to the software. If instrument power  
is interrupted for only a brief time, the instrument will wait  
thirty minutes again. Press the ENTER key to bypass self-  
diagnostic mode.  
Teledyne Analytical Instruments  
Part I 3-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Installation  
Model 2010B  
3-12 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Operation  
4.1 Introduction  
AlthoughtheModel2010Bisusuallyprogrammedtoyourapplicationat  
thefactory,itcanbefurtherconfiguredattheoperatorlevel,oreven,cautious-  
ly,reprogrammed.Dependingonthespecificsoftheapplication,thismight  
includeallorasubsetofthefollowingprocedures:  
Settingsystemparameters:  
Establish a security password, if desired, requiring Operator  
to log in.  
Establish and start an automatic calibration cycle, if desired.  
RoutineOperation:  
Calibratetheinstrument.  
Choose autoranging or select a fixed range of analysis.  
Set alarm setpoints, and modes of alarm operation (latching,  
fail-safe,etc).  
Program/Reprogramtheanalyzer:  
Define new applications.  
Linearize your ranges.  
Ifyouchoosenottousepasswordprotection,thedefaultpasswordis  
automaticallydisplayedonthepasswordscreenwhenyoustartup,andyou  
simplypressEnterforaccesstoallfunctionsoftheanalyzer.  
4.2 Using the Data Entry and Function  
Buttons  
DataEntryButtons:The<>buttonsselectoptionsfromthemenu  
currentlybeingdisplayedontheVFDscreen.Theselectedoptionblinks.  
Whentheselectedoptionincludesamodifiableitem,thearrowbuttons  
canbeusedtoincrementordecrementthatmodifiableitem.  
Teledyne Analytical Instruments  
Part I 4-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
The EnterbuttonisusedtoacceptanynewentriesontheVFDscreen.  
The EscapebuttonisusedtoabortanynewentriesontheVFDscreenthatare  
not yet accepted by use of the Enterbutton.  
Figure4-1showsthehierarchyoffunctionsavailabletotheoperatorviathe  
functionbuttons.Thesixfunctionbuttonsontheanalyzerare:  
Analyze. This is the normal operating mode. The analyzer  
monitors the thermal conductivity of the sample, displays the  
percent or parts-per-million of target gas or contamination, and  
warns of any alarm conditions.  
System. The system function consists of nine subfunctions.  
Four of these are for ordinary setup and operation:  
Setup an Auto-Cal  
Assign Passwords  
Log out to secure system  
InitiateaSelf-Test  
Three of the subfunctions do auxiliary tasks:  
Checking model and software version  
Displaymoresubfunctions  
Two of these are for programming/reprogramming the analyzer:  
Define gas applications and ranges (Refer to programming  
section, or contact factory.)  
Use the Curve Algorithm to linearize output. (Refer to  
programmingsection, orcontactfactory.)  
Zero. Used to set up a zero calibration.  
Span. Used to set up a span calibration.  
Alarms. Used to set the alarm setpoints and determine whether  
each alarm will be active or defeated, HI or LO acting, latching,  
and/orfail-safe.  
Range. Used to set up three analysis ranges that can be  
switched automatically with autoranging or used as individual  
fixedranges.  
Anyfunctioncanbeselectedatanytimebypressingtheappropriatebutton  
(unlesspasswordrestrictionsapply).Theorderaspresentedinthismanualis  
appropriateforaninitialsetup.  
Eachofthesefunctionsisdescribedingreaterdetailinthefollowingproce-  
dures.TheVFDscreentextthataccompanieseachoperationisreproduced,at  
theappropriatepointintheprocedure,inaMonospacedtypestyle.Push-  
buttonnamesareprintedinObliquetype.  
4-2 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
System  
Set LCD  
CONTRAST  
Contrast Function is DISABLED  
Contrast  
(Refer to Section 1.6)  
Span/Zero  
Off/On  
Span/Zero  
Timing  
Span/Zero  
Off/On  
AUTO-CAL  
PASSWORD  
LOGOUT  
Enter  
Password  
Change Yes  
Yes/No  
Change  
Password  
Verify  
Password  
Secure Sys &  
Analyze Only  
MORE  
Show Model  
and Version  
MODEL  
Select  
Range  
Define  
Appl/Range  
APPLICATION  
SELF-TEST  
Self-Test in  
Progress  
Slef-Test  
Results  
Verify  
Points  
Ver  
Enter  
Input/Output  
Values  
Gas Use  
Range  
Select  
Verify/Setup  
Select  
Range  
Man  
ALGORITHM  
Enter  
Enter  
Auto/Manual  
Linearity Cal  
Set  
Select Linrty  
Span Values  
Auto  
Zero in  
Progress  
Auto/Manual  
Zero Select  
Zero  
Auto/Manual  
Span Select  
Span Value  
Set  
Span in  
Progress  
Span  
Select  
Range  
Gas Use  
Range  
% / ppm  
Select  
Setpoints &  
Attributes  
Alarms  
Man  
Define  
Range  
Select  
Range  
Auto/Manual  
Range Adj  
Range  
Gas  
Application  
Auto  
Analyze  
Sample  
Analyze  
Figure 4-1: Hierarchy of Functions and Subfunctions  
Teledyne Analytical Instruments  
Part I 4-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
4.3 The System Function  
ThesubfunctionsoftheSystemfunctionaredescribedbelow.Specific  
proceduresfortheirusefollowthedescriptions:  
AUTO-CAL: Used to define an automatic calibration sequence  
and/or start an AUTO-CAL.  
PWD: Security can be established by choosing a 3 digit  
password (PWD) from the standard ASCII character set. Once a  
unique password is assigned and activated, the operator MUST  
enter the UNIQUE password to gain access to setup functions  
which alter the instrument's operation.  
LOGOUT: Logging out prevents an unauthorized tampering  
withanalyzersettings.  
MORE: Select and enter MORE to get a new screen with  
additionalsubfunctionslisted.  
MODEL: Displays Manufacturer, Model, and Software Version  
ofinstrument.  
APPLICATION: A restricted function, not generally accessed by  
theenduser.Usedtodefineuptothreeanalysisrangesanda  
calibrationrange(includingimpurity,background,lowendofrange,  
highendofrange,and%orppmunits).  
SELF-TEST: The instrument performs a self-diagnostic test to  
check the integrity of the power supply, output boards, sensor  
cell,andpreamplifiers.  
ALGORITHM: A restricted function, not generally accessed by the  
end user. Used to linearize the output for the range of interest.  
4-4 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
4.3.1 Setting the Display  
ContrastFunctionisDISABLED  
(Refer to Section 1.6)  
IfyoucannotreadanythingontheVFDafterfirstpoweringup:  
1. ObserveLEDreadout.  
a. If LED meter reads 8.8.8.8.8., go to step 3.  
b. If LED meter displays anything else, go to step 2.  
2. Press I/O button twice to turn Analyzer OFF and ON again. LED  
meter should now read 8.8.8.8.8.. Go to step 3.  
4.3.2 Setting up an AUTO-CAL  
Whenproperautomaticvalvingisconnected(seechapter3,installation),  
theAnalyzercancycleitselfthroughasequenceofstepsthatautomaticallyzero  
andspantheinstrument.  
Note: Before setting up an AUTO-CAL, be sure you understand the  
Zero and Span functions as described in section 4.4, and  
follow the precautions given there.  
Note: If you require highly accurate AUTO-CAL timing, use external  
AUTO-CAL control where possible. The internal clock in the  
Model 2010B is accurate to 2-3 %. Accordingly, internally  
scheduled calibrations can vary 2-3 % per day.  
Note: If all your ranges are for the same gas application, then AUTO-  
CAL will calibrate whichever range you are in at the scheduled  
time for automatic calibration.  
Note: If your ranges are configured for different applications, then  
AUTO-CAL will calibrate all of the ranges simultaneously (by  
calibrating the Cal Range).  
To setup an Auto–Cal cycle:  
ChooseSystemfromtheFunctionbuttons.TheVFD willdisplayfive  
subfunctions.  
CONTRAST AUTO—CAL  
PWD LOGOUT MORE  
Contrast Function is DISABLED  
(Refer to Section 1.6)  
Teledyne Analytical Instruments  
Part I 4-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
Use < > arrows to blink AUTO—CAL, and press Enter. A new screen for  
ZERO/SPAN set appears.  
ZERO in Ød Øh off  
SPANin Ød Øhoff  
Press < > arrows to blink ZERO(or SPAN), then press Enter again.  
(You won’t be able to set OFF to ON if a zero interval is entered.) A Span  
Every ... (or Zero Every ...) screen appears.  
Zeroschedule:OFF  
Day: ØdHour: Øh  
Use arrows to set an interval value, then use < >arrowstomovetothe  
start-timevalue.Usearrowstosetastart-timevalue.  
To turn ON the SPAN and/or ZERO cycles (to activate AUTO–CAL):  
Press System again, choose AUTO–CAL, and press Enter again. When the  
ZERO/SPAN values screen appears, use the < > arrows to blink the ZERO  
(or SPAN) and press Enter to go to the next screen. Use < > to select OFF/  
ONfield. Use arrows to set the OFF/ONfield to ON. You can now turn  
thesefieldsONbecausethereisanonzerospanintervaldefined.  
4.3.3 Password Protection  
Beforeauniquepasswordisassigned,thesystemassignsTAIbydefault.  
Thispasswordwillbedisplayedautomatically.Theoperatorjustpressesthe  
Enterkeytobeallowedtotalaccesstotheinstrument’sfeatures.  
Ifapasswordisassigned,thensettingthefollowingsystemparameterscan  
bedoneonlyafterthepasswordisentered:alarmsetpoints,assigninganew  
password,range/applicationselections,andcurvealgorithmlinearization.  
(APPLICATIONandALGORITHMarecoveredintheprogrammingsection.)  
However,theinstrumentcanstillbeusedforanalysisorforinitiatingaself-test  
withoutenteringthepassword.Todefeatsecuritythepasswordmustbe  
changed back to TAI.  
NOTE: If you use password security, it is advisable to keep a copy of  
the password in a separate, safe location.  
4-6 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
4.3.3.1  
EnteringthePassword  
Toinstallanewpasswordorchangeapreviouslyinstalledpassword,you  
mustkeyinandENTERtheoldpasswordfirst.Ifthedefaultpasswordisin  
effect, pressingtheENTERbuttonwillenterthedefaultTAIpasswordforyou.  
Press System to enter the System mode.  
ContrastFunctionisDISABLED  
CONTRAST AUTO—CAL  
PWD LOGOUT MORE  
(Refer to Section 1.6)  
Use the < > arrow keys to scroll the blinking over to PWD, and press  
Enterto select the password function. Either the default TAI password or AAA  
placeholdersforanexistingpasswordwillappearonscreendependingon  
whetherornotapasswordhasbeenpreviouslyinstalled.  
Enterpassword:  
TAI  
or  
Enterpassword:  
AAA  
Thescreenpromptsyoutoenterthecurrentpassword.Ifyouarenotusing  
password protection, press Enter to accept TAI as the default password. If a  
passwordhasbeenpreviouslyinstalled,enterthepasswordusingthe<>arrow  
keystoscrollbackandforthbetweenletters, andthearrowkeystochange  
the letters to the proper password. Press Enter to enter the password.  
Inafewseconds,youwillbegiventheopportunitytochangethispass-  
word or keep it and go on.  
ChangePassword?  
<ENT>=Yes <ESC>=No  
Press Escape to move on, or proceed as in Changing the Password,  
below.  
4.3.3.2  
InstallingorChangingthePassword  
Ifyouwanttoinstallapassword,orchangeanexistingpassword,proceed  
asaboveinEnteringthePassword. Whenyouaregiventheopportunityto  
changethepassword:  
ChangePassword?  
<ENT>=Yes <ESC>=No  
PressEntertochange the password(eitherthe defaultTAI or the previ-  
ouslyassignedpassword),orpressEscapetokeeptheexistingpasswordand  
moveon.  
Teledyne Analytical Instruments  
Part I 4-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
IfyouchoseEntertochangethepassword,thepasswordassignment  
screenappears.  
Selectnewpassword  
TAI  
Enter the password using the < > arrowkeys tomove backand forth  
betweentheexistingpasswordletters,andthearrowkeystochangethe  
letterstothenewpassword.Thefullsetof94charactersavailableforpassword  
useareshowninthetablebelow.  
Characters Available for Password Definition:  
A
K
U
_
i
s
}
)
3
=
B
L
V
`
j
t
*
4
>
C
M
W
a
k
u
!
+
5
?
D
N
X
b
l
v
"
'
6
@
E
O
Y
c
m
w
#
-
7
F
P
Z
d
n
x
$
.
8
G
Q
[
e
o
y
%
/
9
H
R
¥
f
p
z
&
0
:
I
S
]
g
q
{
'
1
;
J
T
^
h
r
|
(
2
<
Whenyouhavefinishedtypingthenewpassword,pressEnter.Averifica-  
tionscreenappears.Thescreenwillpromptyoutoretypeyourpasswordfor  
verification.  
EnterPWDToVerify:  
AAA  
Use the arrow keys to retype your password and press Enter when  
finished.Yourpasswordwillbestoredinthemicroprocessorandthesystemwill  
immediatelyswitchtotheAnalyzescreen,andyounowhaveaccesstoall  
instrumentfunctions.  
Ifallalarmsaredefeated, theAnalyzescreenappearsas:  
1.95 %H2inN2  
nR1: Ø1ØAnlz  
Ifanalarmistripped,thesecondlinewillchangetoshowwhichalarmitis:  
1.95 %H2inN2  
AL—1  
4-8 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
NOTE:If you log off the system using the LOGOUT function in the  
system menu, you will now be required to re-enter the pass-  
word to gain access to Alarm, and Range functions.  
4.3.4 Logging Out  
TheLOGOUTfunction providesaconvenientmeansofleavingtheanalyzer  
inapasswordprotectedmodewithouthavingtoshuttheinstrumentoff.By  
enteringLOGOUT,youeffectivelylogofftheinstrumentleavingthesystem  
protectedagainstuseuntilthepasswordisreentered.Tologout,pressthe  
SystembuttontoentertheSystemfunction.  
CONTRAST AUTO—CAL  
PWD LOGOUT MORE  
ContrastFunctionisDISABLED  
(Refer to Section 1.6)  
Use the < > arrow keys to position the blinking over the LOGOUT func-  
tion,andpressEntertoLogout.Thescreenwilldisplaythemessage:  
Protecteduntil  
passwordentered  
4.3.5 System Self-Diagnostic Test  
TheModel2010Bhasabuilt-inself-diagnostictestingroutine.Pre-pro-  
grammedsignalsaresentthroughthepowersupply,outputboard,preamp  
boardandsensorcircuit.Thereturnsignalisanalyzed,andattheendofthetest  
thestatusofeachfunctionisdisplayedonthescreen, eitherasOK orasa  
number between 1 and 1024. (See System Self Diagnostic Test in chapter 5  
fornumbercode.)Ifanyofthefunctionsfails,theSystemAlarmistripped.  
Note: The sensor will always show failed unless identical gas is  
present in both channels at the time of the SELF-TEST.  
Theselfdiagnosticsarerunautomaticallybytheanalyzerwheneverthe  
instrumentisturnedon,butthetestcanalsoberunbytheoperatoratwill.To  
initiateaselfdiagnostictestduringoperation:  
PresstheSystembuttontostarttheSystemfunction.  
CONTRAST AUTO—CAL  
PWD LOGOUT MORE  
Contrast Function is DISABLED  
(Refer to Section 1.6)  
Use the < > arrow keys to blink MORE, then press Enter.  
MODEL APPLICATION  
SELF—TEST ALGORITHM  
Use the < > arrow keys again to move the blinking to the SELF–TEST  
andpressEnter.Thescreenwillfollowtherunningofthediagnostic.  
Teledyne Analytical Instruments  
Part I 4-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
RUNNING DIAGNOSTIC  
TestingPreampCell  
Whenthetestingiscomplete,theresultsaredisplayed.  
Power:OK Analog:OK  
Cell: 2 Preamp:3  
ThemoduleisfunctioningproperlyifitisfollowedbyOK.Anumber  
indicatesaprobleminaspecificareaoftheinstrument.RefertoChapter5  
MaintenanceandTroubleshootingfornumber-codeinformation.Theresults  
screenalternatesforatimewith:  
PressAnyKey  
ToContinue...  
ThentheanalyzerreturnstotheinitialSystemscreen.  
4.3.6 The Model Screen  
Move the < > arrow key to MORE and press Enter. With MODEL  
blinking,pressEnter.Thescreendisplaysthemanufacturer,model,andsoft-  
wareversioninformation.  
4.3.7 Checking Linearity with ALGORITHM  
From the System Function screen, select ALGORITHM, and press  
Enter.  
selrngtosetalgo:  
—> Ø1 Ø2 Ø3 <—  
Use the < > keys to select the range: 01, 02, or 03. Then press Enter.  
Gas Use: H2 — N2  
Range:  
Ø
— 10%  
PressEnteragain.  
Algorithmsetup:  
VERIFY SETUP  
Select and EnterVERIFY to check whether the linearization has been  
accomplishedsatisfactorily.  
Dpt INPUT OUTPUT  
Ø Ø.ØØ Ø.ØØ  
4-10 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Theleftmostdigit(underDpt)isthenumberofthedatapointbeingmoni-  
tored.Usethekeystoselectthesuccessivepoints.  
TheINPUTvalueistheinputtothelinearizer.Itisthesimulatedoutputof  
theanalyzer.Youdonotneedtoactuallyflowgas.  
The OUTPUT value is the output of the linearizer. It should be the ACTU-  
ALconcentrationofthespangasbeingsimulated.  
IftheOUTPUTvalueshownisnotcorrect,thelinearizationmustbe  
corrected. PressESCAPEtoreturntothepreviousscreen. SelectandEnter  
SET UP to Calibration Mode screen.  
Selectalgorithm  
mode : AUTO  
There are two ways to linearize: AUTO and MANUAL: The auto mode  
requiresasmanycalibrationgasesastherewillbecorrectionpointsalongthe  
curve.Theuserdecidesonthenumberofpoints,basedontheprecisionre-  
quired.  
Themanualmodeonlyrequiresenteringthevaluesforeachcorrection  
pointintothemicroprocessorviathefrontpanelbuttons.Again,thenumberof  
pointsrequiredisdeterminedbytheuser.  
4.4 The Zero and Span Functions  
(1)TheModel2010Bcanhaveasmanyasthreeanalysisrangesplusa  
specialcalibrationrange(CalRange);andtheanalysisranges,ifmorethanone,  
maybeprogrammedforseparateoridenticalgasapplications.  
(2)Ifallrangesareforthesameapplication,thenyouwillnotneedtheCal  
Range.Calibratinganyoneoftherangeswillautomaticallycalibratetheothers.  
(3)If:a)eachrangeisprogrammedforadifferentgasapplication,b)your  
sensorcalibrationhasdriftedlessthan10 %,andc)yourCalRangewascali-  
bratedalongwithyourotherrangeswhenlastcalibrated,thenyoucanusethe  
CalRangetocalibrateallapplicationsrangesatonce.  
IfyourModel2010Banalyzerfitstheparagraph(3)description,above,  
usetheCalRange.Ifyouranalyzerhasdriftedmorethan10 %,calibrateeach  
rangeindividually.  
CAUTION: Always allow 4-5 hours warm-up time before calibrat-  
ing, if your analyzer has been disconnected from its  
power source. This does not apply if the analyzer  
was plugged in but was in STANDBY.  
Teledyne Analytical Instruments  
Part I 4-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
Theanalyzeriscalibratedusingreference,zero,andspangases.Gas  
requirementsarecoveredindetailinchapter3,section3.4GasConnections.  
Checkthatcalibrationgasesareconnectedtotheanalyzeraccordingtothe  
instructionsinsection3.4,observingalltheprescribedprecautions.  
Note: Shut off the gas pressure before connecting it to the analyzer,  
and be sure to limit pressure to 40 psig or less when turning it  
back on.  
Readjustthegaspressureintotheanalyzeruntiltheflowratethroughthe  
sensorsettlesbetween50to200cc/min(approximately0.1to0.4scfh).  
Note: Always keep the zero calibration gases flow as close to the  
flowrate of sample gas as possible  
4.4.1 Zero Cal  
TheZerobuttononthefrontpanelisusedtoenterthezerocalibration  
function.Zerocalibrationcanbeperformedineithertheautomaticormanual  
mode.  
CAUTION: If you are zeroing the Cal Range by itself (multiple  
application analyzers only), use manual mode  
zeroing.  
If you want to calibrate ALL of the ranges at once  
(multiple application analyzers only), use auto mode  
zeroing in the Cal Range.  
Make sure the zero gas is flowing to the instrument. If you get a CELL  
CANNOT BE BALANCED message while zeroing skip to section 4.4.1.3.  
4.4.1.1  
AutoModeZeroing  
Observe the precautions in sections 4.4 and 4.4.1, above. PressZeroto  
enterthezerofunctionmode.Thescreenallowsyoutoselectwhetherthezero  
calibrationistobeperformedautomaticallyormanually. Usethearrow  
keys to toggle between AUTO and MAN zero settling. Stop when AUTO  
appears,blinking,onthedisplay.  
Selectzero  
mode: AUTO  
PressEntertobeginzeroing.  
####.##% H2 — N2  
Slope=#.### C—Zero  
4-12 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Thebeginningzerolevelisshownintheupperleftcornerofthedisplay.As  
thezeroreadingsettles,thescreendisplaysandupdatesinformationonSlope=  
inpercent/second(unlesstheSlopestartswithintheacceptablezerorangeand  
doesnotneedtosettlefurther).Thesystemfirstdoesacoursezero,shownin  
thelowerrightcornerofthescreenasC—Zero, for3 min, andthendoesafine  
zero,anddisplaysF—Zero,for3 min.  
Then,andwheneverSlopeislessthan0.01foratleast3min,insteadof  
Slope you will see a countdown: 9 Left, 8 Left, and so fourth. These are  
softwarestepsinthezeroingprocessthatthesystemmustcomplete,AFTER  
settling, before it can go back toAnalyze. Software zero is indicated by S–  
Zero in the lower right corner.  
####.## % H2N2  
4Left=#.### S—Zero  
Thezeroingprocesswillautomaticallyconcludewhentheoutputiswithin  
theacceptablerangeforagoodzero.Thentheanalyzerautomaticallyreturnsto  
the Analyze mode.  
4.4.1.2  
ManualModeZeroing  
PressZerotoentertheZerofunction.Thescreenthatappearsallowsyou  
toselectbetweenautomaticormanualzerocalibration.Usethekeysto  
toggle between AUTO and MAN zero settling. Stop when MANUAL appears,  
blinking,onthedisplay.  
Selectzero  
mode: MANUAL  
PressEntertobeginthezerocalibration.Afterafewsecondsthefirstof  
threezeroingscreensappears.Thenumberintheupperlefthandcorneristhe  
first-stagezerooffset.Themicroprocessorsamplestheoutputatapredeter-  
minedrate.  
####.## % H2N2  
Zeroadj:2048C—Zero  
TheanalyzergoesthroughC–Zero,F–Zero,andS–Zero.DuringC–Zero  
and F–Zero, use the keys to adjust displayed Zero adj: value as close as  
possibletozero.Then,pressEnter.  
S–Zerostarts.DuringS–Zero,theMicrocontrollertakescontrolasinAuto  
ModeZeroing,above.Itcalculatesthedifferencesbetweensuccessivesam-  
plingsanddisplaystherateofchangeasSlope= avalueinpartspermillionper  
second(ppm/s).  
Teledyne Analytical Instruments  
Part I 4-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
####.## % H2 — N2  
Slope=#.### S—Zero  
Generally, you have a good zero when Slope is less than 0.05 ppm/s for  
about30seconds.  
Oncezerosettlingcompletes,theinformationisstoredintheanalyzer’s  
memory,andtheinstrumentautomaticallyreturnstotheAnalyzemode.  
4.4.1.3 Cell Failure  
Cellfailureinthe2010Bisusuallyassociatedwithinabilitytozerothe  
instrumentwithareasonablevoltagedifferentialacrosstheWheatstonebridge.If  
thisshouldeverhappen,the2010Bsystemalarmtrips,andtheVFDdisplaysa  
failuremessage.  
Cellcannotbebalanced  
Checkyourzerogas  
Beforereplacingthesensor:  
a. Check your zero gas to make sure it is within specifications.  
b. Check for leaks downstream from the sensor, where contamina-  
tion may be leaking into the system.  
c. Check flowmeter to ensure that the flow is no more than  
200SCCM  
d. Checktemperaturecontrollerboard.  
e. Check gas temperature.  
Ifnoneoftheaboveasindicated, thesensormayneedtobereplaced.  
Checkwarranty,andcontactAnalyticalInstrumentsCustomerService.  
4.4.2 Span Cal  
TheSpanbuttononthefrontpanelisusedtospancalibratetheanalyzer.  
Spancalibrationcanbeperformedineithertheautomaticormanualmode.  
CAUTION: If you are spanning the Cal Range by itself (multiple  
application analyzers only), use manual mode  
zeroing.  
If you want to calibrate ALL of the ranges at once  
(multiple application analyzers only), use auto mode  
spanning in the Cal Range.  
Makesurethespangasisflowingtotheinstrument.  
4-14 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
4.4.2.1  
AutoModeSpanning  
Observeallprecautionsinsections4.4and4.4.2, above. PressSpanto  
enterthespanfunction.Thescreenthatappearsallowsyoutoselectwhetherthe  
spancalibrationistobeperformedautomaticallyormanually. Usethe∇  
arrow keys to toggle between AUTO and MAN span settling. Stop when  
AUTO appears, blinking, onthedisplay.  
Selectspan  
mode: AUTO  
PressEnterto move to the next screen.  
SpanVal: 2Ø.ØØ%  
<ENT>Tobeginspan  
Usethe<>arrowkeystotogglebetweenthespanconcentrationvalue  
andtheunitsfield(%/ppm).Usethearrowkeyschangethevalueand/orthe  
units,asnecessary.Whenyouhavesettheconcentrationofthespangasyouare  
using,pressEntertobegintheSpancalibration.  
####.##% H2 — N2  
Slope=#.### Span  
Thebeginningspanvalueisshownintheupperleftcornerofthedisplay.  
Asthespanreadingsettles,thescreendisplaysandupdatesinformationon  
Slope.Spanningautomaticallyendswhenthespanoutputcorresponds,within  
tolerance,tothevalueofthespangasconcentration.Thentheinstrumentauto-  
maticallyreturnstotheanalyzemode.  
4.4.2.2  
ManualModeSpanning  
PressSpantostarttheSpanfunction.Thescreenthatappearsallows  
youtoselectwhetherthespancalibrationistobeperformedautomaticallyor  
manually.  
Selectspan  
mode: MANUAL  
Use the keys to toggle between AUTO and MAN span settling. Stop  
when MAN appears, blinking, on the display. Press Enterto move to the next  
screen.  
SpanVal: 2Ø.ØØ%  
<ENT>Tobeginspan  
Usethe<>arrowkeystotogglebetweenthespanconcentrationvalue  
andtheunitsfield(%/ppm).Usethearrowkeyschangethevalueand/orthe  
Teledyne Analytical Instruments  
Part I 4-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
units,asnecessary.Whenyouhavesettheconcentrationofthespangasyouare  
using,pressEntertobegintheSpancalibration.  
PressEntertoenterthespanvalueintothesystemandbeginthespan  
calibration.  
Oncethespanhasbegun,themicroprocessorsamplestheoutputata  
predeterminedrate.Itcalculatesthedifferencebetweensuccessivesamplings  
anddisplaysthisdifferenceasSlopeonthescreen.Ittakesseveralsecondsfor  
thefirstSlopevaluetodisplay.SlopeindicatesrateofchangeoftheSpan  
reading.Itisasensitiveindicatorofstability.  
####.##% H2N2  
Slope=#.### Span  
WhentheSpanvaluedisplayedonthescreenissufficientlystable,press  
Enter.(Generally,whentheSpanreadingchangesby1 %orlessoftherange  
beingcalibratedforaperiodoftenminutesitissufficientlystable.)OnceEnter  
ispressed,theSpanreadingchangestothecorrectvalue.Theinstrumentthen  
automaticallyenterstheAnalyzefunction.  
4.5 The Alarms Function  
TheModel2010Bisequippedwith6fullyadjustablesetpointsconcentra-  
tionwithtwoalarmsandasystemfailurealarmrelay.Eachalarmrelayhasaset  
offormC"contactsratedfor3amperesresistiveloadat250V ac.SeeFigure  
inChapter3,Installationand/ortheInterconnectionDiagramincludedatthe  
backofthismanualforrelayterminalconnections.  
Thesystemfailurealarmhasafixedconfigurationdescribedinchapter3  
Installation.  
Theconcentrationalarmscanbeconfiguredfromthefrontpanelaseither  
high or low alarms by the operator. The alarm modes can be set as latching or  
non-latching, and either failsafe or non-failsafe, or, they can be defeated  
altogether.Thesetpointsforthealarmsarealsoestablishedusingthisfunction.  
Decidehowyouralarmsshouldbeconfigured.Thechoicewilldepend  
uponyourprocess.Considerthefollowingfourpoints:  
1. Which if any of the alarms are to be high alarms and which if any  
are to be low alarms?  
Setting an alarm as HIGH triggers the alarm when the  
contaminant concentration rises above the setpoint. Setting an  
alarm as LOW triggers the alarm when the contaminant  
concentration falls below the setpoint.  
Decide whether you want the alarms to be set as:  
4-16 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Both high (high and high-high) alarms, or  
One high and one low alarm, or  
Both low (low and low-low) alarms.  
2. Are either or both of the alarms to be configured as failsafe?  
In failsafe mode, the alarm relay de-energizes in an alarm  
condition. For non-failsafe operation, the relay is energized in an  
alarm condition. You can set either or both of the concentration  
alarms to operate in failsafe or non-failsafe mode.  
3. Are either of the alarms to be latching?  
In latching mode, once the alarm or alarms trigger, they will  
remain in the alarm mode even if process conditions revert back  
to non-alarm conditions. This mode requires an alarm to be  
recognized before it can be reset. In the non-latching mode, the  
alarm status will terminate when process conditions revert to non-  
alarmconditions.  
4. Are either of the alarms to be defeated?  
The defeat alarm mode is incorporated into the alarm circuit so  
that maintenance can be performed under conditions which  
would normally activate the alarms.  
The defeat function can also be used to reset a latched alarm.  
(See procedures, below.)  
Ifyouareusingpasswordprotection,youwillneedtoenteryourpassword  
toaccessthealarmfunctions.Followtheinstructionsinsection4.3.3toenter  
yourpassword.Onceyouhaveclearancetoproceed,entertheAlarmfunction.  
Note: If all ranges are for the same application, setting one of them  
will automaticaly set the others.  
PresstheAlarmbuttononthefrontpaneltoentertheAlarmfunction.  
Makesurethat01isblinking.  
Selrngtosetalm:  
—> Ø1 Ø2 Ø3 <—  
SetuptheRange1alarmbymovingtheblinkingoverto01usingthe<>  
arrowkeys.ThenpressEnter.Checkthegasapplicationandrangelimitsas  
displayedonthescreen.  
Gas use: H2 — N2  
Range: 010%  
Pressenteragaintosetthealarmsetpoints.  
Sel%/ppmalmtoset  
AL1—PPM AL2—PPM  
Teledyne Analytical Instruments  
Part I 4-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
Use the keys to choose between % or ppm units. Then press Enter to  
movetothenextscreen.  
AL1: 1ØØØ ppm HI  
Dft:N Fs:N Ltch:N  
Fiveparameterscanbechangedonthisscreen:  
Value of the alarm setpoint, AL1: ####  
Out-of-range direction, HI or LO  
Defeated? Dft:Y/N (Yes/No)  
Failsafe? Fs:Y/N (Yes/No)  
Latching? Ltch:Y/N (Yes/No).  
To define the setpoint, use the < > arrow keys to move the  
blinking over to AL1: ####. Then use the arrow keys to  
change the number. Holding down the key speeds up the  
incrementingordecrementing.  
To set the other parameters use the < > arrow keys to move the  
blinking over to the desired parameter. Then use the arrow  
keys to change the parameter.  
Once the parameters for alarm 1 have been set, press Alarms  
again, and repeat this procedure for alarm 2 (AL2).  
To reset a latched alarm, go to Dft– and then press either two  
times or two times. (Toggle it to Y and then back to N.)  
–OR –  
Go to Ltch– and then press either two times or two times.  
(Toggle it to N and back to Y.)  
4.6 The Range Select Function  
TheRangefunctionallowsyoutomanuallyselecttheconcentration range  
of analysis (MANUAL), or to select automatic range switching (AUTO).  
In the MANUAL screen, you are further allowed to define the high and low  
(concentration)limitsofeachRange,andselectasingle,fixedrangetorun.  
CAUTION: If this is a linearized application, the new range must  
be within the limits previously programmed using the  
System function, if linearization is to apply through-  
out the range. Furthermore, if the limits are too small  
a part (approx 10 % or less) of the originally linear-  
ized range, the linearization will be compromised.  
4-18 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
IntheAUTOscreen,youarefurtherallowedtoselectwhichgasapplication  
(PREVIOUSLYdefinedinSystemfunction)torun.  
4.6.1 Manual (Select/Define Range) Screen  
TheManualrange-switchingmodeallowsyoutoselectasingle,fixed  
analysisrange.Itthenallowsyoutoredefinetheupperandlowerlimits,forthe  
range.  
PressRangekeytostarttheRangefunction.  
Selectrange  
mode: MANUAL  
Note: If all three ranges are currently defined for different applica-  
tion gases, then the above screen does not display (because  
mode must be manual). Instead, the VFD goes directly to the  
following screen.  
If above screen displays, use the arrow keys to Select MANUAL, and  
pressEnter.  
Selectrangetorun  
—> Ø1 Ø2 Ø3 CAL<—  
Use the < > keys to select the range: 01, 02, 03, or CAL. Then press  
Enter.  
Gasuse: H2N2  
Range: Ø — 10 %  
Usethe <> keystotogglebetweentheRange:low-endfieldandthe  
Range:high-endfield.Usethekeystochangethevaluesofthefields.  
PressEscapetoreturntothepreviousscreentoselectordefineanother  
range.  
PressEntertoreturnthetotheAnalyzefunction.  
4.6.2 Auto (Single Application) Screen  
Autorangingwillautomaticalysettotheapplicationthathasatleasttwo  
rangesetupwiththesamegases.  
Intheautorangingmode,themicroprocessorautomaticallyrespondsto  
concentrationchangesbyswitchingrangesforoptimumreadoutsensitivity.Ifthe  
upperlimitoftheoperatingrangeisreached,theinstrumentautomaticallyshifts  
tothenexthigherrange.Iftheconcentrationfallstobelow85%offullscaleof  
Teledyne Analytical Instruments  
Part I 4-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
thenextlowerrange,theinstrumentswitchestothelowerrange.Acorrespond-  
ingshiftintheDCconcentrationoutput,andintherangeIDoutputs,willbe  
noticed.  
Theautorangingfeaturecanbeoverriddensothatanalogoutputstays ona  
fixedrangeregardlessofthecontaminantconcentrationdetected.Iftheconcen-  
trationexceedstheupperlimitoftherange,theDCoutputwillsaturateat1V dc  
(20mAatthecurrentoutput).  
However,thedigitalreadoutandtheRS-232outputoftheconcentration  
areunaffectedbythefixedrange.Theycontinuetoreadbeyondthefull-scale  
settinguntilamplifiersaturationisreached.Belowamplifiersaturation,the  
overrangereadingsareaccurateUNLESStheapplicationuseslinearizationover  
theselectedrange.  
TheconcentrationrangescanberedefinedusingtheRangefunction  
Manualscreen,andtheapplicationgasescanberedefinedusingtheSystem  
function,iftheyarenotalreadydefinedasnecessary.  
CAUTION: Redefining applications or ranges might require  
relinearizationand/orrecalibration.  
Tosetupautomaticranging:  
PressRangekeytostarttheRangefunction.  
Selectrange  
mode : AUTO  
Note: If all three ranges are currently defined for different applica-  
tion gases, then the above screen does not display (because  
mode must be manual).  
If above screen displays, use the arrow keys to Select AUTO, and  
press Enter.  
PressEscapetoreturntothepreviousAnalyzeFunction.  
4-20 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
4.6.3 Precautions  
Part I: Control Unit  
TheModel2010Ballowsagreatdealofflexibilityinchoosingrangesfor  
automaticrangeswitching.However,therearesomepitfallsthataretobe  
avoided.  
Rangesthatworkwelltogetherare:  
Ranges that have the same lower limits but upper limits that differ  
by approximately an order of magnitude  
Ranges whose upper limits coincide with the lower limits of the  
next higher range  
Ranges where there is a gap between the upper limit of the range  
and the lower limit of the next higher range.  
Rangeschemesthataretobeavoidedinclude:  
Ranges that overlap  
Ranges whose limits are entirely within the span of an adjoining  
range.  
Figure4-2illustratestheseschemesgraphically.  
Figure 4-2: Examples of Autoranging Schemes  
4.7 The Analyze Function  
Normally,allofthefunctionsautomaticallyswitchbacktotheAnalyze  
functionwhentheyhavecompletedtheirassignedoperations.Pressingthe  
Teledyne Analytical Instruments  
Part I 4-21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
EscapebuttoninmanycasesalsoswitchestheanalyzerbacktotheAnalyze  
function.Alternatively,youcanpresstheAnalyzebuttonatanytimetoreturn  
toanalyzingyoursample.  
TheAnalyzefunctionscreenshowstheimpurityconcentrationandthe  
applicationgasesinthefirstline,andtherangeinthesecondline.Inthelower  
rightcorner,theabbreviationAnlzindicatesthattheanalyzerisintheAnalyze  
mode.Ifthereisan*beforetheAnlz,itindicatesthattherangeislinearized.  
1.95 % H2 N2  
nR1:Ø10 *Anlz  
nindicatesnoninvertingrange  
iindicatesinvertingrange  
Iftheconcentrationdetectedisoverrange,thefirstlineofthedisplayblinks  
continuously.  
4.8 Programming  
CAUTION: The programming functions of the Set Range and  
Curve Algorithm screens are configured at the facto-  
ry to the users application specification. These func-  
tions should only be reprogrammed by trained,  
qualified personnel.  
Toprogram,youmust:  
1. Enter the password, if you are using the analyzer’s password  
protectioncapability.  
2. Connect a computer or computer terminal capable of sending an  
RS-232 signal to the analyzer RS-232 connector. (See chapter 3  
Installation for details). Send the rp command to the analyzer.  
OR  
For software 1.1.4 or later, turn the instrument off and back on.  
While on the introduction screen hold the Analyze key for at least  
fifteen seconds. Press the Enter key twice to return to the  
Analyze mode.  
3. Press the System button to start the System function.  
CONTRAST AUTO—CAL  
PWD LOGOUT MORE  
ContrastFunctionisDISABLED  
(Refer to Section 1.6)  
Use the < > arrow keys to blink MORE, then press Enter.  
4-22 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
MODEL APPLICATION  
SELF_TEST ALGORITHM  
NowyouwillbeabletoselecttheAPPLICATIONandALGORITHM  
set-upfunctions.  
4.8.1 The Set Range Screen  
TheSetRangescreenallowsreprogrammingofthethreeanalysisranges  
andthecalibrationrange(includingimpuritygas,backgroundgas,lowendof  
range,highendofrange,and%orppmunits).Originalprogrammingisusually  
doneatthefactoryaccordingtothecustomer’sapplication.Itmustbedone  
throughtheRS-232portusingacomputerrunningaterminalemulationprogram.  
Note: It is important to distinguish between this System programming  
subfunction and the Range button function, which is an operator  
control. The Set Range Screen of the System function allows the user  
to DEFINE the upper and lower limits of a range AND the application  
of the range. The Range button function only allows the user to  
select or define the limits, or to select the application, but not  
to define the application.  
NormallytheModel2010Bisfactorysettodefaulttomanualrange  
selection,unlessitisorderedasasingle-applicationmultiple-rangeunit(inwhich  
caseitdefaultstoautoranging).Ineithercase,autorangingormanualrange  
selectioncanbeprogrammedbytheuser.  
Intheautorangingmode,themicroprocessorautomaticallyrespondsto  
concentrationchangesbyswitchingrangesforoptimumreadoutsensitivity.Ifthe  
upperlimitoftheoperatingrangeisreached,theinstrumentautomaticallyshifts  
tothenexthigherrange.Iftheconcentrationfallstobelow85%offullscaleof  
thenextlowerrange,theinstrumentswitchestothelowerrange.Acorrespond-  
ingshiftintheDCconcentrationoutput,andintherangeIDoutputs,willbe  
noticed.  
Theautorangingfeaturecanbeoverriddensothatanalogoutputstays ona  
fixedrangeregardlessofthecontaminantconcentrationdetected.Iftheconcen-  
trationexceedstheupperlimitoftherange,theDCoutputwillsaturateat1V dc  
(20mAatthecurrentoutput).  
However,thedigitalreadoutandtheRS-232outputoftheconcentration  
areunaffectedbythefixedrange.Theycontinuetoreadbeyondthefull-scale  
settinguntilamplifiersaturationisreached.Belowamplifiersaturation,the  
overrangereadingsareaccurateUNLESStheapplicationuseslinearizationover  
theselectedrange.  
Toprogramtheranges,youmustfirstperformthefourstepsindicatedat  
thebeginningofsection4.8Programming.Youwillthenbeinthesecond  
Systemmenuscreen.  
Teledyne Analytical Instruments  
Part I 4-23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
MODEL APPLICATION  
SELF_TEST ALGORITHM  
Use the < > arrow keys again to move the blinking to APPLICATION  
and press Enter.  
Selrngtosetappl:  
—> Ø1 Ø2 Ø3 CAL <—  
Usethearrowkeystoincrement/decrementtherangenumberto01,  
02, 03, or CAL, and press Enter.  
Imp: H2 Bck: N2  
FR:Ø TO:1Ø %  
Usethe<>arrowkeystomovetoImp:(impurity),Bck:(background),  
FR:(from—lowerendofrange),TO:(to—upperendofrange),andPPMor%.  
Usethearrowkeystoincrementtherespectiveparametersasdesired.  
Press Enter to accept the values and return to Analyze mode. (See note  
below.)Repeatforeachrangeyouwanttoset.  
Note: The ranges must be increasing from low to high, for example,  
if Range 1 is set to 0–10 % and Range 2 is set to 0–100 %, then  
Range 3 cannot be set to 0–50 % since that makes Range 3  
lower than Range 2.  
Ranges,alarms,andspansarealwayssetineitherpercentorppmunits,as  
selectedbytheoperator,eventhoughallconcentration-dataoutputschange  
fromppmtopercentwhentheconcentrationisabove9999 ppm.  
Note: When performing analysis on a fixed range, if the concentra-  
tion rises above the upper limit as established by the operator  
for that particular range, the output saturates at 1 V dc (or 20  
mA). However, the digital readout and the RS-232 output  
continue to read regardless of the analog output range.  
Toendthesession:  
IfstartedwiththeRS-232,send:  
st<enter>  
st<enter>  
totheanalyzerfromthecomputer.  
Ifstartedthroughthefrontpanel,turntheinstrumentoffandbackon.  
Press the Enter key twice to return to the Analyze mode.  
4-24 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
4.8.2 The Curve Algorithm Screen  
TheCurveAlgorithmisalinearizationmethod.Itprovidesfrom1to9  
intermediatepointsbetweentheZEROandSPANvalues,whichcanbenormal-  
izedduringcalibration,toensureastraight-lineinput/outputtransferfunction  
throughtheanalyzer.  
Eachrangeislinearizedindividually,asnecessary,sinceeachrangewill  
usuallyhaveatotallydifferentlinearizationrequirement. Beforesettingthe  
algorithmcurve,eachrangemustbeZeroedandSpanned.  
Tolinearizetheranges,youmustfirstperformthefourstepsindicatedat  
thebeginningofsection4.8Programming.Youwillthenbeinthesecond  
Systemmenuscreen.  
MODEL APPLICATION  
SELF_TEST ALGORITHM  
4.8.2.1 Checkingthelinearization  
From the System Function screen, select ALGORITHM, and press  
Enter.  
Selrngsetalgo  
—> Ø1 Ø2 Ø3 <—  
Use the < > keys to select the range: 01, 02, or 03. Then press Enter.  
Gasuse:H2 N2  
Range: Ø-10 %  
PressEnteragain.  
Algorithmsetup:  
VERIFY SETUP  
Select and EnterVERIFY to check whether the linearization has been  
accomplishedsatisfactorily.  
Dpt INPUT OUTPUT  
Ø Ø.ØØ Ø.ØØ  
Theleftmostdigit(underDpt)isthenumberofthedatapointbeingmoni-  
tored.Usethekeystoselectthesuccessivepoints.  
TheINPUTvalueistheinputtothelinearizer.Itisthesimulatedoutputof  
theanalyzer.Youdonotneedtoactuallyflowgas.  
The OUTPUT value is the output of the linearizer. It should be the ACTU-  
ALconcentrationofthespangasbeingsimulated.  
Teledyne Analytical Instruments  
Part I 4-25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
IftheOUTPUTvalueshownisnotcorrect,thelinearizationmustbecorrect-  
ed. Press ESCAPE to return to the previous screen. Select and Enter SET UP  
toCalibrationModescreen.  
Select algorithm  
mode : AUTO  
There are two ways to linearize: AUTO and MANUAL: The auto mode  
requiresasmanycalibrationgasesastherewillbecorrectionpointsalongthe  
curve.Theuserdecidesonthenumberofpoints,basedontheprecisionre-  
quired.  
Themanualmodeonlyrequiresenteringthevaluesforeachcorrection  
pointintothemicroprocessorviathefrontpanelbuttons.Again,thenumberof  
pointsrequiredisdeterminedbytheuser.  
Note: Before performing section 4.8.2 or 4.8.2.3, you must check to  
ensure that your calibration gases or points are between low  
end and high end of the range setup. All correction points  
must be between Zero and Span concentrations. Do not enter  
Zero and Span points as part of the correction.  
4.8.2.2 Manual Mode Linearization  
Tolinearizemanually,youmusthavepreviousknowledgeofthenonlinear  
thermal-conductivitycharacteristicsofyourgases.Youenterthevalueofthe  
differentialbetweentheactualconcentrationandtheapparentconcentration  
(analyzeroutput).AnalyticalInstrumentshastabulardataofthistypeforalarge  
numberofgases,whichitmakesavailabletocustomersonrequest.SeeAppen-  
dixfororderinginformation.Toenterdata:  
From the System Functions Screen—  
1. Use < > to select ALGORITHM , and Enter.  
2. Select and Enter SETUP.  
3. Enter MANUAL from the Calibration Mode Select screen.  
Dpt INPUT OUTPUT  
Ø Ø.ØØ Ø.ØØ  
Thedataentryscreenresemblestheverifyscreen,butthegasvaluescan  
bemodifiedandthedata-pointnumbercannot. Usethe<>keystotoggle  
between the INPUT and OUTPUT fields. Use the keys to set the value for  
thelowestconcentrationintothefirstpoint.ThenpressEnter.  
4-26 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Aftereachpointisentered,thedata-pointnumberincrementstothenext  
point.Movingfromthelowesttothehighestconcentration,usethekeysto  
setthepropervaluesateachpoint.  
Dpt INPUT  
0 Ø.ØØ Ø.ØØ  
OUTPUT  
Repeattheaboveprocedureforeachofthedatapointsyouaresetting(up  
toninepoints:0-8).Setthepointsinunitincrements.Donotskipnumbers.The  
linearizerwillautomaticallyadjustforthenumberofpointsentered.  
When you are done, Press ESCAPE. The message, Completed. Wait  
for calculation, appears briefly, and then the main System screen returns.  
Toendthesession:  
IfstartedwiththeRS-232,send:  
st<enter>  
st<enter>  
totheanalyzerfromthecomputer.  
Ifstartedthroughthefrontpanel,turntheinstrumentoffandbackon.  
Press the Enter key twice to return to the Analyze mode.  
4.8.2.3 Auto Mode Linearization  
TolinearizeintheAutoMode,youmusthaveonhandaseparatecalibra-  
tiongasforeachofthedatapointsyouaregoinguseinyourlinearization.First,  
theanalyzeriszeroedandspannedasusual.Then,eachspecialcalibrationgas,  
foreachoftheintermediatecalibrationpoints,isflowed,inturn,throughthe  
sensor.Aseachgasflows,thedifferentialvalueforthatintermediatecalibration  
pointisenteredfromthefrontpaneloftheanalyzer.  
Note: The span gas use to span the analyzer must be >90% of the  
range being analyzed.  
Beforestartinglinearization,performastandardcalibration.Seesection  
4.4. Toenterdata:  
From the System Functions screen—  
1. Use < > to select ALGORITHM , and Enter.  
2. Select and Enter SETUP.  
3. Enter AUTO from the Calibration Mode Select screen.  
TheAutoLinearizeModedataentryscreenappears.  
1.95% H2 N2  
Input(Ø):2.00  
Teledyne Analytical Instruments  
Part I 4-27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
5. Use the keys to set the proper value of calibration gas, and  
Enter. Repeat this step for each cal-point number as it appears in  
the Input (x) parentheses.  
6. Repeat step 5 for each of the special calibration gases, from the  
lowest to the highest concentrations. Press Escape when done.  
Toendthesession:  
IfstartedwiththeRS-232,send:  
st<enter>  
st<enter>  
totheanalyzerfromthecomputer.  
Ifstartedthroughthefrontpanel,turntheinstrumentoffandbackon.  
Press the Enter key twice to return to the Analyze mode.  
4.9 Special Function Setup  
4.9.1 Output Signal Reversal  
Someapplicationsrequireareversaloftheoutputsignalsinorderforthe  
4-20mAand0-1VDCoutputsignalstocorrespondwiththelowandhighend  
oftheconcentrationrange. Forexample,ifanapplicationinvolvestheanalysis  
of85%oxygeninabackgroundofargonbymeasuringthethermalconductivity  
ofthebinarygas,theanalyzerwouldnormallybesetupsothatthe100%  
oxygen(0%argon)concentrationwouldcorrespondtothezerolevel(4mA0V)  
oftheoutputsignal. Then,85%oxygen(15%argon)wouldcorrespondto  
20mA(1V)inthesignaloutput.  
Itmaybeconvenientfortheusertohavetheoutputsreversedsothatthe  
85-100%oxygenleveloutputsa4-20mA(0-1V)signalrespectively. Thiscan  
beaccomplishedbyreversingthedatainputtothecustomsettings. Notall  
applicationswillrequireareversingfunction,however,ifthisisdesirable,itmust  
bespecifiedatthetimeofpurchaseoralternatively,bysubstitutingalinearizing  
PCboardwiththereversalinformationcontainedtherein. Contactthefactory  
forfurtherinformation.  
4.9.2 Special - Inverting Output  
NOTE: If the unit has a range or ranges that specified >0 unit setup  
for inverting.  
Thestepsare:  
1.  
Press RANGE key.  
4-28 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
2.  
UseLEFT/RIGHTkeytomovetotherangethatisspeci-  
fiedasinvertingoutput.  
3.  
4.  
PressandholdDOWNkeyforapproximately5to7seconds.  
PressENTER key.  
NOTE: If the inverting has been setup, “i” shall display on the left  
bottom corner. Otherwise, the left bottom corner display ”n”.  
Ifmorethatonerangeisspecifiedaninvertingoutput,repeatsteps1  
to 4.  
4.9.3 Special - Polarity Coding  
NOTE: This setup will be identified only when performing GAS TEST  
or calculation. Formula 1 will determine which range(s)  
willrequired polarity coding.  
Iftheanalyzerindicatesanegativeconcentrationwithproperspangasor  
formula1indicatesanegativevalue,settheS1accordinglytothetablebelow:  
Close S1-5 range 1  
Close S1-6 range 2  
Close S1-7 range 3  
CloseS1-8calrange  
PressI/Otorestartthesystem.  
4.9.4 Special - Nonlinear Application Gain Preset  
NOTE: This section apply during GAS TEST routine for the unit that  
has more than one range install with nonlinear output applica-  
tion.  
Thestepsareasfollows:  
1.  
2.  
Setunitrangetolowestrangereading.  
Usingthecomputergeneratedsettingsforthecontroller,  
adjustthecontrollersettingsforthemaximumspangas  
output.  
3.  
PressSPANkey. SelectAUTOmodeandsetupthesetting  
to span level. Press ENTER key to span.  
4.  
5.  
Setrangeswitchtonextrange.  
Press SPAN key.  
Teledyne Analytical Instruments  
Part I 4-29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Operation  
Model 2010B  
6.  
Press and hold the RIGHT key for approximate 5 to 7  
seconds.  
7.  
SelectAUTOandsetthereadingtospangaslevel. Press  
ENTERkey.  
Repeat steps 1 to 7 if more than two ranges need to be setup.  
4-30 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Maintenance  
5.1 Routine Maintenance  
Aside from normal cleaning and checking for leaks at the gas connec-  
tions, routine maintenance is limited to replacing fuses, and recalibration. For  
recalibration, see Section 4.4 Calibration.  
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS  
MANUAL.  
5.2 System Self Diagnostic Test  
1. Press the System button to enter the system mode.  
2. Use the < > arrow keys to move to More, and press Enter.  
3. Use the < > arrow keys to move to Self-Test, and press Enter.  
The following failure codes apply:  
Table 5-1: Self Test Failure Codes  
Power  
0
1
2
3
OK  
5 V Failure  
15 V Failure  
Both Failed  
Analog  
0
1
2
3
OK  
DAC A (0–1 V Concentration)  
DAC B (0–1 V Range ID)  
Both Failed  
(Continued)  
Part I 5-1  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5 Maintenance  
Model 2010B  
Preamp  
0
OK - > 0 means one or more gain have a  
high offset, e.g. 204 means highest  
gain has an offset. This error is  
common on sealed air cells.  
Cell  
0
OK  
1
Failed (open filament, short to ground, no  
power.)  
2
Unbalance(deteriorationoffilaments,blocked  
tube)  
5.3 VFD Display  
NOTE:  
When a vacuum Fluorescent Display is used, It will not  
require contrast adjustment.  
If you cannot read anything on the VFD, especially after first powering  
up, check that VFD cable is not loose.  
5.4 Fuse Replacement  
The 2010B requires two 5 x 20 mm, 1 A, T type (Slow Blow) fuses.  
The fuses are located inside the enclosure on the Electrical Connector Panel,  
as shown in Figure 5-1. To replace a fuse:  
1. Disconnect the Unit from its power source.  
2. Place a small screwdriver in the notch in the fuse holder cap,  
push in, and rotate 1/4 turn. The cap will pop out a few  
millimeters. Pull out the fuse holder cap and fuse, as shown in  
Figure 5-1.  
5-2 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part I: Control Unit  
Figure 5-1: Removing Fuse Cap and Fuse from Holder  
3. Replace fuse by reversing process in step 1.  
Remove Power to the instrument before changing fuses.  
5.5 Major Internal Components  
The Interconnection Panel and the Front Panel PCBs are accessed by  
unlatching and swinging open the front door, as described earlier.  
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS  
MANUAL.  
The 2010B Control Unit contains the following major components:  
Power Supply  
PreampandMotherboardwithMicrocontroller  
Display Board and Displays  
5 digit LED meter  
2 line, 20 character, alphanumeric, VFD  
Rear Panel Board.  
Part I 5-3  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5 Maintenance  
Model 2010B  
5.6 Cleaning  
If instrument is unmounted at time of cleaning, disconnect the instru-  
ment from the power source. Close and latch the front-panel access door.  
Clean outside surfaces with a soft cloth dampened slightly with plain clean  
water. Do not use any harsh solvents such as paint thinner or benzene.  
For mounted instruments, DO NOT wipe the front panel while the  
instrument is controlling your process. Clean the front panel as prescribed in  
the above paragraph.  
5.7 Phone Numbers  
CustomerService: (626) 934-1673  
Environmental Health and Safety: (626) 961-9221, Extension 1592  
Fax: (626) 961-2538  
EMERGENCY ONLY: (24-hour pager) 1-800-759-7243  
PIN # 1858192  
5-4 Part I  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATING INSTRUCTIONS  
Model 2010B  
Thermal Conductivity Analyzer  
Part II: Analysis Unit  
NEC Type  
Part Number D-70089  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 2010B Thermal Conductivity Analyzer  
Table of Contents  
1 Introduction  
1.1 Overview........................................................................ 1-1  
1.2 Connections................................................................... 1-1  
1.3 Electrical Connector Panel ............................................ 1-2  
2 Installation  
2.1 Unpacking the Analysis Unit.......................................... 2-1  
2.2 Mounting the Analysis Unit ............................................ 2-1  
2.3 Sample System Design ................................................. 2-2  
2.4 Pressure and Flowrate Regulation ................................ 2-3  
2.5 VENT Exhaust ............................................................... 2-3  
2.6 SAMPLE Gas................................................................. 2-4  
2.7 REFERENCE Gas......................................................... 2-4  
2.8 ZERO Gas ..................................................................... 2-5  
2.9 SPAN Gas...................................................................... 2-5  
2.10 Electrical Connector Panel ............................................ 2-5  
2.11 Signal Wiring Recommendations................................... 2-7  
2.12 Testing the System......................................................... 2-7  
3 Operation  
3.1 System Self Diagnostic Test........................................... 3-1  
3.2 Cell Failure Checks ....................................................... 3-2  
4 Maintenance  
4.1 Routine Maintenance..................................................... 4-1  
4.2 Major Components......................................................... 4-1  
4.3 Fuse Replacement......................................................... 4-2  
4.4 System Self Diagnostic Test........................................... 4-3  
4.5 Cell, Heater, and/or Thermistor Replacement ................ 4-4  
4.5.1 Removing the Cell Compartment ........................ 4-4  
4.5.2 Removing and Replacing the Cell Block............. 4-5  
4.5.3 Removing the Heater and/or Thermocouple........ 4-6  
4.5.4 Replacing the Heater and/or Thermister .............. 4-7  
ii: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part II: Analysis Unit  
Introduction  
1.1 Overview  
The Analytical Instruments Model 2010B Analysis Unit is a versatile  
remotely controlled instrument for measuring a component gas in a back-  
ground gas, or in a specific mixture of background gases.  
Part 1 of this manual covers the Control Unit. Part II, this part, covers  
the Model 2010B NEC type explosion proof Analysis Unit only.  
1.2 Connections  
The standard 2010 Analysis Unit is housed in a NEC type housing and  
includes all gas connections. Figure 1-1 is a cutaway illustration of the  
Analysis Unit showing the Gas Connections. The gas connectors are de-  
scribed briefly here and in detail in the Installation chapter of this manual.  
NOTE: The version of the 2010 Analysis Unit depicted has no options.  
The gas connections found on the 2010B Analysis Unit depend on the  
options selected. Refer to the outline diagram in the rear of this manual for  
the actual configuration. These options may include gas panels with  
flowmeters and auto-cal or manual valves as well as flame arrestors. If you  
have purchased a unit with a sealed air option, there will not be any refer-  
ence gas connections or flowmeter. If you have not purchased a unit with  
gas panel, provision must be made to provide zero and span gas to the  
sample in port.  
SAMPLE IN This is the gas input from the sample stream.  
This gas connection should have a flowmeter  
with a range of 50-200 cc/min.  
Teledyne Analytical Instruments  
Part II: 1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 2010B  
Figure 1-1: Outline Diagram of 2010B Analysis Unit  
REFERENCE IN This is the gas input from the flowing  
reference gas source. This gas connector should  
have a flowmeter with a range of 20-100 cc/  
min.  
SPAN/ZERO These gas inputs have the same requirements as  
the sample in connection.  
VENTS  
Sample and reference gas vents must be re-  
turned to areas of equal pressure.  
The actual tubing size and connection type will depend on actual  
options selected. See outline diagram in the rear of the manual for piping  
requirements.  
1.3 Electrical Connector Panel  
Figure 1-2 shows the internal Electrical Connector Panel. Cables enter  
the housing through access ports (visible in Figure 1-1), and connect to  
terminals inside the housing. The connectors and controls are described  
briefly here. They are described in detail in the Installation, Operation, and  
Maintenancechapters, asappropriate.  
1-2: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part II: Analysis Unit  
Figure 1-2: Electrical Connector/Control Panel  
Power In  
Power input terminals for electric heater.  
Requires 110 or 220 V ac, depending on  
position of the Voltage Selector switch. Use  
50/60 Hz.  
CAUTION: Check the position of the Voltage Selector switch  
BEFORE applying power to the Power Input termi-  
nals.  
VoltageSelector Power input selector switch for electric heater.  
Adjusts input requirement for 110 or  
220 V ac, depending on available source  
voltage. Use 50/60 Hz.  
Fuses  
1.6 A, 250 V, T type, European size  
5 × 20 mm fuses. Fuse 2 is on the neutral side  
of the line. Fuse 1 is on the hot side of the  
line.  
Solenoid Valves Terminalsthatprovideallelectricalintercon-  
nections from the Control Unit to the gas  
controlvalves.  
Sensor Signal  
Terminals that provide connections from the  
T/C sensor to the Control Unit.  
Teledyne Analytical Instruments  
Part II: 1-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Introduction  
Model 2010B  
1-4: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part II: Analysis Unit  
Installation  
Installation of the Model 2010B Analyzer includes:  
1. Unpacking, mounting, and interconnecting the Control Unit and  
the Analysis Unit  
2. Making gas connections to the system  
3. Making electrical connections to the system  
4. Testing the system.  
2.1 Unpacking the Analysis Unit  
The analyzer is shipped with all materials needed to install and prepare the  
system for operation. Carefully unpack the Analysis Unit and inspect it for  
damage. Immediately report any damage to the shipping agent.  
2.2 Mounting the Analysis Unit  
The Model 2010B Analysis Unit is for use in Class 1, Division 1,  
Groups C and D, hazardous environments (group B available). The actual  
class is limited by the flame arrestor option.  
The standard model is designed for bulkhead mounting. Overall dimen-  
sions of the Analysis Unit will vary depending on options. The maximum  
footprint will be 19″ × 12, and maximum height 9.4. An Outline Drawing  
at the back of this manual, gives the correct mounting dimensions for your  
unit.  
Note: The housing, including the cover, protrudes 8.6 to 9.4inches  
from the base on which it is mounted. Enough clearance is  
required in front of the cover to allow the cover to be re-  
moved.  
Figure 3-1 is a view with the cover removed showing the external Gas  
Connector Panel and the internal Electrical Connector Panel.  
Teledyne Analytical Instruments  
Part II: 2-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2 Installation  
Model 2010B  
Figure 3-1: 2010 Analysis Unit with Auto-Cal & Gas Panel options  
2-3 Sample System Design  
Gas Connector and Selector Panels for specific applications are avail-  
able at additional cost . These panels are designed to substitute a standard  
frontpanel.  
For those customers wishing to incorporate their own sample system,  
electronic input/output ports are provided on the rear panel for the operation  
of solenoid valves under the complete control of the Model 2010B electron-  
ics. See section 3.3. The recommended system piping schematic is included  
among the drawings at the rear of the manual  
For best results, use the recommended piping system. Select a  
flowmeter that can resolve 0.08 scfh (40-50 cc/min) for the reference path of  
the analyzer, and select a flowmeter that can resolve 0.3 scfh (150 cc/min)  
for the sample path of the analyzer.  
2-2: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part II: Analysis Unit  
NOTE:  
NOTE:  
The sample-line pressure regulator should be installed as  
close to the sample point as possible to minimize sample-  
line lag time.  
An additional option is available for SEALED reference  
application. This option would not have the reference Gas  
Flow Meter, Piping and Fittings.  
2-4 Pressure and Flowrate Regulation  
Appropriate pressure reducing regulators must be installed at all gas  
supply sources. To minimize flowrate adjustments the pressure regulators on  
the supporting gas supply cylinders should be adjusted to provide the same  
output pressure as the sample line regulator.  
The gas pressure at the IN input should be reasonably well regulated.  
Pressures between 5 and 50 psig are acceptable (10 psig is normal) as long  
as the pressure, once established, will keep the flow constant during analy-  
sis and within an acceptable range (between 0.1 and 0.4 scfh—See Note).  
Note: Gases lighter than air have a flowrate higher than indicated on  
the flowmeter, while gases heavier than air have a flowrate  
lower than indicated. Values can range from one half to twice  
the indicated flowrate.  
For example: For hydrogen or helium, set the flowrate to 0.1  
scfh (50 cc/min). For carbon dioxide or argon, set the flowrate  
to 0.4 scfh (200 cc/min).  
When installing pressure regulators on supply cylinders, crack the  
cylinder valves so that gas is flowing during installation. This will eliminate  
the most common cause of standardization-gas contamination: air trapped  
during assembly diffusing back into the cylinder. This procedure is particu-  
larly important in applications where impurity content of 1 to 2 % is the  
range of interest.  
2-5 VENT Exhaust  
There are two separate VENT fittings—one for the sample gas and one  
for the reference gas. Use 1/4 inch tubing for both sample and reference vents  
to minimize back pressure from restricted flow.  
Exhaust connections must be consistent with the hazard level of the  
constituent gases. Check local, state, and federal laws, and ensure that the  
exhaust stream vents to an appropriately controlled area if required. If not  
vented to the same area, both VENT lines must vent to areas with equal  
Teledyne Analytical Instruments  
Part II: 2-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2 Installation  
Model 2010B  
ambient pressures, and pressures must vary no more than the normal baro-  
metricchanges.  
2-6 SAMPLE Gas  
In the standard model, sample and calibration gases are introduced  
through the SAMPLE fitting. The gases must be Tee'd into the Sample inlet  
withappropriatevalves.  
The gas pressure in should be well regulated. The sample line pressure  
regulator should be installed as close to the sample line as possible to mini-  
mize sample line lag time.  
If greater flow is required for improved response time, install a bypass  
in the sampling system upstream of the analyzer input.  
2-7 REFERENCE Gas  
A gas of fixed composition is needed as a reference to which the  
sample gas will be compared. The reference gas is normally selected to  
represent the main background gas of the analysis.  
For most applications, a constant supply of reference gas flowing at the  
same rate as the sample is required for best results. However, in many cases  
the flow of reference gas can be slowed to about 0.08 scfh (40 cc/min) with  
goodresults.  
For some applications, an optional sealed air reference is installed. In  
sealed-reference sensors the reference side of the detector cell is filled with  
air and sealed. This eliminates the need to have reference gas constantly  
passing through the cell.  
NOTE:For instruments equipped with the optional sealed air refer-  
ence, there is no REFERENCE inlet or reference VENT port.  
It is highly recommended that the same cylinder of gas be used for both  
the REFERENCE gas and the ZERO gas.  
Pressure, flow, and safety considerations are the same as prescribed for  
the SAMPLE gas, above.  
2-4: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
2-8 ZERO Gas  
Part II: Analysis Unit  
For the ZERO gas, a supply of the background gas, usually containing  
none of the impurity, is required to zero the analyzer during calibration. For  
suppressed zero ranges the zero gas must contain the low-end concentration  
of the impurity.  
NOTE:Because most cylinder gases are between 99.95 and 99.98%  
pure, it is highly recommended that the same cylinder of gas  
be used for both REFERENCE and ZERO gas.  
NOTE:It is essential to the accuracy of the analyzer that the purity of  
the zero gas be known. Otherwise, when the zero control is  
adjusted during zero standardization, the reading will indicate  
the impurity content of the zero gas, rather than zero.  
2-9 SPAN Gas  
For the SPAN gas, a supply of the background gas containing 70-  
100 % of the component of interest is required as a minimum.  
Note: If your analyzer range is set for inverting output, your zero gas  
will be at 100% of the range interest, and span will be 70 to  
100% of the low end range.  
If linearization is required, intermediate concentrations of the target gas  
in the background gas may be necessary. From one to nine separate span  
gases may be used, depending on the desired precision of the linearization.  
See chapter 4, Operation.  
2-10 Electrical Connector Panel  
All electrical connections are made on the internal Electrical Connector  
Panel, inside the explosion-proof enclosure, illustrated in Figure 3-3. The  
signals are described in the following paragraphs. Wire size and length are  
given in the Drawings section at the back of this manual. To access the  
Panel, remove the explosion-proof cover as described in chapter 5, Mainte-  
nance. NEVER OPEN THE COVER IN A HAZARDOUS ATMO-  
SPHERE. THE AREA MUST BE DECLARED TEMPORARILY SAFE  
BY THE PROPER AUTHORITY FIRST.  
Teledyne Analytical Instruments  
Part II: 2-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2 Installation  
Model 2010B  
Figure 3-4: Control Unit (CU) to Analysis Unit (AU) Connector  
If you use your own gas control valves, use the interconnect diagram in  
Figure 3-4 for the valves. The sensor and thermistor remain connected as in  
Figure 3-5, above. (See drawing D-73170 for wire recommendations.)  
Figure 3-5: Remote Probe Connector Pinouts  
The voltage from the solenoid outputs is nominally 0 V for the OFF  
and 15 V dc for the ON conditions. The maximum combined current that  
can be pulled from these output lines is 100 mA. (If two lines are ON at the  
same time, each must be limited to 50 mA, etc.)  
If more current and/or a different voltage is required, use a relay, power  
amplifier, or other matching circuitry to provide the actual driving current.  
Note that each individual line has a series FET with a nominal ON resistance  
of 5 ohms (9 ohms worst case). This can limit the obtainable voltage, de-  
pending on the load impedance applied. See Figure 3-6.  
2-6: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part II: Analysis Unit  
Figure 3-6: FET Series Resistance  
2-11 Signal Wiring Recommendations  
The signal leads are the bias and sensor leads ref fig. 3.4, and reference  
D-73170interconnectiondiagram. Themaximuminterconnectiondistances  
between the 2010 control unit and the analysis unit is limited by the sensor  
signal requirements. The sensor has two limiting factors. The first is series  
resistance which is limited to a maximum of 2.5 ohms. Refer to the intercon-  
nection diagram for wire gages and distances which will meet this require-  
ment.  
The second issue is one of signal quality. At greater distances prob-  
lems associated with signal degradation increases. Signal degradation prob-  
lems have two major issues. The first is associated with the local installation  
environment. The number, proximity, and power associated with local  
sources of EMI radiation can not be accounted for by TAI. These issues can  
only be controlled by the end user. The other factor is the amplitude of the  
signals generated by the thermal conductivity cell. This signal level is a  
function of the gas composition of the gases involved. This signal level can  
vary from micro-volts to mili-volts dependent on application. Consult TAI  
for recommendations based on sample stream composition. Signal quality  
and operating distances can be increased by the uses of better quality  
shielded low loss cable. The cable configuration should be twisted pair with  
double shielding for best performance. The signal cable should not be run  
with AC power or digital signals.  
2-12 Testing the System  
After The Control Unit and the Analysis Unit are both installed and  
interconnected, and the system gas and electrical connections are complete,  
the system is ready to test. Before plugging either of the units into their  
respectivepowersources:  
Teledyne Analytical Instruments  
Part II: 2-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2 Installation  
Model 2010B  
Check the integrity and accuracy of the gas connections. Make  
sure there are no leaks.  
Check the integrity and accuracy of the electrical connections.  
Make sure there are no exposed conductors  
Check that sample pressure is between 3 and 40 psig, according  
to the requirements of your process.  
Power up the system, and test it as follows:  
1. Repeat the Self-Diagnostic Test as described in Part I, chapter 4,  
section 4.3.5.  
2-8: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part II: Analysis Unit  
Operation  
3.1 System Self Diagnostic Test  
The self diagnostics are run automatically by the analyzer whenever the  
instrument is turned on, but the test can also be run by the operator at will.  
During the test, internal signals are sent through the power supply, output  
board and sensor circuit automatically. The return signal is analyzed, and at  
the end of the test the status of each function is displayed on the screen,  
either as OK or as a number between 1 and 3. (See Table 4-1 for number  
code.)  
Note: Remote Probe connector must be connected to the Analysis  
Unit, or sensor circuit will not be properly checked.  
Instructions for running self diagnostics are repeated here for your  
convenience:  
1. Press the System button to enter the system mode.  
2. Use the < > arrow keys to move to More, and press Enter.  
3. Use the < > arrow keys to move to Self-Test, and press Enter.  
During preamp testing there is a countdown in the lower right corner of  
the screen. When the testing is complete, the results are displayed.  
Power:OK Analog:OK  
Preamp:3  
The following failure codes apply:  
Table 4-1: Self Test Failure Codes  
Power  
0
1
OK  
5 V Failure  
Teledyne Analytical Instruments  
Part II: 3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3 Operation  
Model 2010B  
2
3
15 V Failure  
Both Failed  
Analog  
0
1
2
3
OK  
DAC A (0–1 V Concentration)  
DAC B (0–1 V Range ID)  
Both Failed  
Preamp  
OK - > 0 means gains of the amplifier have a  
high offset, e.g. 204 means the highest  
0
gain has a high offset it is common on  
sealedairreference.  
The results screen alternates for a time with:  
Press Any Key  
ToContinue...  
Then the analyzer returns to the initial System screen.  
3.2 Cell Failure Checks  
Cell failure is covered in detail in Part I: Control Units, section 4.4.1.3,  
Cell Failure. Cell replacement is covered Part II: Analysis Units chapter 5,  
Maintenance.  
3-2: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part II: Analysis Unit  
Maintenance  
4.1 Routine Maintenance  
Aside from normal cleaning and checking for leaks at the gas connec-  
tions,routinemaintenanceislimitedtorecalibration.  
Self-diagnostic testing of the system and fuse replacement in the Con-  
trol Unit are covered in Part I, chapter 5 of this manual. For recalibration, see  
Part I, section 4.4 Calibration.  
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS  
MANUAL.  
4.2 Major Components  
The internal components are accessed by rotating the explosion-proof  
housing cover counterclockwise several turns until free. See Figure 4-1,  
below.  
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS  
MANUAL.  
The 2010B Analysis Unit contains the following major components:  
TC Cell  
Optional Auto-Cal Solenoid Assy.  
HeaterAssembly  
ElectricalConnectorPanel  
Gas Connector Panel (external)  
See the drawings in the Drawings section in back of this manual  
fordetails.  
Teledyne Analytical Instruments  
Part II: 4-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Maintenance  
Model 2010B  
Figure 4-1: Major Components  
4.3 Fuse Replacement  
The 2010B Analysis Unit requires two 5 x 20 mm, 1.6 A, T type  
(Slow Blow) fuses. The fuses are located inside the explosion proof housing  
on the Electrical Connector Panel, as shown in Figure 4-2. To replace a fuse:  
1. Disconnect the Unit from its power source.  
2. Place a small screwdriver in the notch in the fuse holder cap,  
push in, and rotate 1/4 turn. The cap will pop out a few  
millimeters. Pull out the fuse holder cap and fuse, as shown in  
Figure 4-2.  
Figure 4-2: Removing Fuse Cap and Fuse from Holder  
4-2: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part II: Analysis Unit  
3. Replace fuse by reversing process in step 1.  
4.4 System Self Diagnostic Test  
1. Press the System button to enter the system mode.  
2. Use the < > arrow keys to move to More, and press Enter.  
3. Use the < > arrow keys to move to Self-Test, and press Enter.  
4. Observe the error-code readings on the VFD Display screen, and  
check Table 4-1, below, to interpret the codes.  
Table 4-1: Self Test Failure Codes  
Power  
0
1
2
3
OK  
5 V Failure  
15 V Failure  
Both Failed  
Analog  
0
1
2
3
OK  
DAC A (0–1 V Concentration)  
DAC B (0–1 V Range ID)  
Both Failed  
Preamp  
OK - > 0 means gains of the amplifier have a  
0
high offset, e.g. 204 means the highest  
gain has a high offset it is common on  
sealedairreference.  
Teledyne Analytical Instruments  
Part II: 4-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Maintenance  
Model 2010B  
4.5 Cell, Heater, and/or Thermistor  
Replacement  
The Thermal Conductivity Cell, with its Heater and Thermistor, is  
mounted inside the insulated cell compartment, inside the analysis unit. To  
remove these components, you must first remove screw on the cover, and  
thecustomerinterface.  
4.5.1 Removing the Cell Compartment  
WARNING: IF THE MODEL 2010B ANALYZER HAS BEEN USED  
WITH TOXIC GASES, FLUSH IT THOROUGHLY  
BEFORE PERFORMING THIS PROCEDURE.  
WARNING: DISCONNECT ALL POWER TO THE MODEL 2010B  
BEFORE PERFORMING THIS PROCEDURE.  
Remove the t/c cell as follows:  
a. Verify that the power has been removed from the analysis unit  
prior to removing cover.  
b. Verify that the analysis unit is free from any toxic gas prior  
disconnectinganyplumbing.  
c. Remove the screw on lid and the six #6 screws securing the  
interfacecoverpanel.  
d. Remove the cover panel and the four 3/4" #6 stands-off securing  
the 1st interface PCB.  
e. Remove the 1st PCB and the four 7/16" #6 stand-off securing the  
2nd interface PCB.  
f. Unplug the heater from J3, and the sensor connectors from the  
sensor connector J4, and J2, on the upper interface PCB.  
Remove the 2nd PCB. Remove the two #10 screws from the  
sensor retaining bracket located at the top of the sensor.  
Disconnect the gas fitting and remove the sensor from the  
analysisunit.  
g. Reassemble in the reverse order, verify that there are no gas leaks  
prior to returning the unit to service.  
4-4: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part II: Analysis Unit  
4.5.2 Removing and Replacing the Cell Block  
a. Refer to Figure 4-3, which illustrates removal of the Cell Block  
from the Cell Compartment. Exploded view is as seen from the  
top of the Cell Block.  
Figure 4-3: Removal of Cell from Cell Housing  
Teledyne Analytical Instruments  
Part II: 4-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Maintenance  
Model 2010B  
b. Remove the two screws holding the front mounting bracket—  
they also hold the Cell Block Cover to the Cell Block—and then  
pull off the cover.  
c. Turn the uncovered Cell Block assembly over so that the bottom  
faces you. The black rectangular block with four screws is the  
Heater Block. Separate the Heater Block from the Cell Block by  
removing the four screws. Leave the Heater Block electrical  
connectionsconnected.  
d. Remove the four screws from each of the black plates that hold  
the Cell. The Cell is sandwiched between the plates. You should  
now be able to slide the Cell free.  
e. Leave the electrical connections connected at the Cell. Unlace the  
cabling, and unplug the grey Cell cable at the Interface PCB  
connector, J3. The Preamplifier PCB can be more easily  
accessed by removing the analyzer's rear panel.  
f. Replace the cell by reversing the above procedure, steps a  
through e.  
g. On analyzers with the sealed air reference option, the thermistor  
is epoxied to the front of the cell block.  
4.5.3 Removing the Heater and/or Thermocouple  
a. Refer to Figure 4-4, which illustrates removal of the Thermistor  
and/or Heater from the Cell Compartment. Exploded view is as  
seen from the bottom of the Cell Block.  
To Interface Board  
Thermistor  
Heater  
Figure 4-4: Removing the Heater and/or Thermocouple  
4-6: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Part II: Analysis Unit  
b. Remove the two screws holding the front mounting bracket—  
they also hold the Cell Block Cover to the Cell Block—and then  
pull off the cover.  
c. Turn the uncovered Cell Block assembly over so that the bottom  
faces you. The black rectangular block with four screws is the  
Heater Block.  
d. The Heater is fastened to the Heater Block by a set screw as well  
as the silicone sealing compound. The Thermistor is fastened  
only by the silicone sealer. (NOTE: On units with sealed air  
reference option, the thermistor is epoxied to the front on the  
cell.)  
(1) To remove the Heater, use a 1/16 Allen wrench to loosen the  
Thermistor set screw. Then, grasp BOTH Heater wires  
firmly, and pull the Heater slowly out of the Heater Block,  
breaking the silicone seal. Do not allow any foreign matter to  
enter the empty duct.  
(2) To remove the Thermistor, grasp BOTH Thermistor wires  
firmly, and pull the Thermistor slowly out of the Heater  
Block, breaking the silicone seal. Do not allow any foreign  
matter to enter the empty duct.  
e. Undo the cable lacing and separate the Heater/Thermistor wires.  
Then, disconnect the wires from the interface PCB.  
4.5.4 Replacing the Heater and/or Thermister  
a. To replace the Heater and/or Thermister, coat the new element  
with silicone sealing compound, and insert it into the duct.  
CAUTION: The larger duct is for the Heater element, and the  
smaller duct is for the Thermocouple.  
b. Enough sealing compound should be on the element to spill over  
and seal around the wire where it enters the duct. Smooth the  
outer seal and remove any excess. (NOTE: On unit with the  
sealed air reference option, the thermistor is expoxied to the  
front of the t/c cell.)  
c. Reassemble the Cell Compartment by reversing the procedure in  
section 4.5.3. Then relace the cabling.  
d. Reinstall the Assembled Cell Compartment by reversing the  
procedure in section 4.5.1. Then reconnect the wires to J2 on the  
interface PCB.  
Teledyne Analytical Instruments  
Part II: 4-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4 Maintenance  
Model 2010B  
4-8: Part II  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATING INSTRUCTIONS  
Models 2010B  
Thermal Conductivity Analyzer  
Appendix  
Bulkhead Mount Control Unit, PN D70845  
NEC Type Analysis Unit, PN D70089  
NEC Type Analysis Unit, PN D70147  
Teledyne Analytical Instrum
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Model 2010B  
Contents  
A-1 Model 2010B Specifications .......................................... A-3  
A-2 Recommended 2-Year Spare Parts List ......................... A-5  
A-3 Drawing List ................................................................... A-6  
A-2  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Appendix  
Appendix  
A-1 Specifications  
Ranges: Three ranges plus a cal range, field selectable  
within limits (application dependent) and Auto  
Ranging  
Display: 2 line by 20 alphanumeric VFD accompanied  
by 5 digit LED display  
Accuracy: ±1% of full scale for most binary mixtures at  
constanttemperature  
±5% of full scale over operating temperature  
range once temperature equilibrium has been  
reached  
Response Time: 90% in less than 50 seconds  
SystemOperating  
Temperature: 32°F to 122°F (0 - 50°C)  
Sensor Type: StandardTCcell(4-filamentdetector)  
SignalOutput: Two 0-1 VDC (concentration and range ID)  
Two 4-20 mADC isolated (concentration and  
range ID)  
Alarm: Twofullyprogrammableconcentrationalarm  
set points and corresponding Form C, 3 amp  
contacts.  
One system failure alarm contact to detect  
power, calibration, zero / span and sensor  
failure.  
Teledyne Analytical Instruments  
A-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Model 2010B  
System Power  
Requirements: Ex-Proof Analysis Unit 110 or 220 VAC, 50-  
60Hz. (InternalSwitchselectable)  
Control Unit: Universal Power 85-250 VAC 47-63Hz.  
CellMaterial: Nickel plated brass block with nickel alloy  
filaments and stainless steel end plates  
O/PInterface: Full duplex RS-232, implement a subset of  
Tracs Command  
Mounting:  
Standard: General purpose NEMA-4 Bulkhead mounted  
Control Unit  
Bulk Head mount Analysis Unit  
Options: General purpose relay rack mounted to con-  
tain either one or two in a 19” relay rack  
mountableplate  
* Other configurations, including a totally  
explosion-proof,areavailable  
Dimensions:  
Control Unit: 15.75”H x 11.75”W x 8.5”D  
Analysis Unit: 13.6”W x 18.1”H x 8.25”D  
Control Unit: General Purpose Flush Panel Mounting  
Analysis Unit: Explosion Proof Enclosure is UL and CSA  
Listed for Class 1, Division 1, Groups B, C,  
D Service. Nema 4/7 rated. (Actual Group  
Rating is dependent on options selected)  
Watted Parts: 316SS (except cell) standard version.  
(Auto-Cal or Gas Panel option include viton  
seals. The standard flowmeter includes viton  
seals, SS valves, glass metering tube in an  
acrylicblock).  
A-4  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Thermal Conductivity Analyzer  
Appendix  
A-2 Recommended 2-Year Spare Parts List  
Qty  
1
Part Number Description  
D67472  
Back Panel Board  
1
C62371B  
C65098A  
C62365D  
D70082  
D67651  
F768  
Front Panel Board  
1
PreamplifierBoard  
1
Main Computer Board  
1
Customer Interface PCB Assy (AU)  
Std. T/C Cell Assy. (Brass + SS)  
Fuse 1.6A (Analysis Unit)  
2010 Front Door w/overlay  
Fuse, 1 A, 250V, 5x20mm, T- Slo-Blo  
Customer Interface PCB (CU)  
1
2
1
D64849B  
F1275  
2
1
D65295B  
_____________________  
* Order one type only: US or European, as appropriate.  
Note: Orders for replacement parts should include the part number  
(if available) and the model and serial number of the instru-  
ment for which the parts are intended.  
Orders should be sent to:  
TELEDYNE Analytical Instruments  
16830 Chestnut Street  
City of Industry, CA 91749-1580  
Phone (626) 934-1500  
Fax (626) 961-2538  
or your local representative.  
Teledyne Analytical Instruments  
A-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix  
Model 2010B  
A-3 Drawing List  
C70381  
D70117  
C70146  
C70335  
D70081  
D-73170  
Outline Drawing (Standard & Auto-Cal option -C)  
PipingDiagram  
Outline Diagram (AU Auto-Cal Valve and Gas Panel -C-L)  
Outline Diagram (gas panel option -L)  
SchematicDiagram, InterfaceBoard  
InterconnectionDiagram  
A-6  
Teledyne Analytical Instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Tascam Car Stereo System DA 98HR User Manual
Technika Flat Panel Television LCD19W 219 User Manual
Texas Instruments Computer Hardware MSP430x11x1 User Manual
Therma Stor Products Group Dehumidifier R175 User Manual
THORENS Turntable TD 2010 User Manual
Toastmaster Waffle Iron TMRWB User Manual
Transition Networks Switch TX12 User Manual
Tripp Lite TV Cables S455 006 User Manual
ViewSonic Digital Photo Frame VFA724W User Manual
VTech Cordless Telephone VT 2420 User Manual