OPERATING INSTRUCTIONS FOR
Model 2010B
Thermal Conductivity Analyzer
Teledyne Analytical Instruments
0.0
AL-1
%
Anlz
2010B
Thermal Conductivity Analyzer
DANGER
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING
SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
P/N M70845
11/22/00
ECO # 00-0517
ONLYAUTHORIZEDPERSONNELSHOULDCONDUCTMAINTENANCEAND/ORSERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/
MANAGER.
Teledyne Analytical Instruments
i
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Table of Contents
Specific Model Information ................................. iv
Part I: Control Unit, Model 2010B .........Part I: 1-1
Part II: Analysis Unit, Model 2010B......Part II: 1-1
Appendix ......................................................... A-1
Teledyne Analytical Instruments
iii
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 2010B
Specific Model Information
The instrument for which this manual was supplied may incorporate one or more
options not supplied in the standard instrument. Commonly available options are listed
below, with check boxes. Any that are incorporated in the instrument for which this
manual is supplied are indicated by a check mark in the box.
Instrument Serial Number: _______________________
Options Included in the Instrument with the Above Serial Number:
❑ C:
Auto Calibration valves (zero/span/sample) built-in gas control
valves are electronically controlled to provide synchronization
with the analyzer’s operations.
❑ G:
❑ H:
❑ K:
Stainless steel cell block with nickel filaments and Stainless
Steel fittings and tubing.
Stainless steel cell block with gold filaments for corrosive gas
streams and Stainless Steel fittings and tubing.
19" Rack Mount option with one or two analyzer Control
Units installed and ready to mount in a standard rack.
❑ K2:
19" Rack Mount option with two Control Units mounted.
❑ K3:
19" Rack Mount option with one Control Unit mounted and a
blank cover installed in the second Control Unit location.
❑ L:
Gas selector panel consisting of sample/ref flow meters and
control valves for metering input of sample/calibrations support
gases.
❑ F:
Flame Arrestors for Class 1, Div. 1, Groups C/D service.
❑ P:
Flame Arrestors for Class 1, Div. 1, Groups C/D service, and
Auto Cal valves option (Ref. C above) and GP use.
❑ Q:
Flame Arrestors for Group B (hydrogen) service, and Auto Cal
valves option (Ref. C above)
❑ O:
Flame Arrestors for Group B (hydrogen)
Sealed Reference Cell
❑ R:
iv
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Table of Contents
1 Introduction
1.1 Overview ........................................................................ 1-1
1.2 Typical Applications ....................................................... 1-2
1.3 Main Features of the Analyzer ....................................... 1-2
1.4 Model Designations ....................................................... 1-3
1.5 Operator Interface (Front Panel) .................................... 1-3
1.6 Recognizing Difference Between LCD & VFD ............... 1-5
1.7 Equipment Interface (Rear Panel).................................. 1-5
1.8 Gas Connections ........................................................... 1-7
2 Operational Theory
2.1 Introduction .................................................................... 2-1
2.2 Sensor Theory ............................................................... 2-1
2.2.1 Sensor Configuration .............................................. 2-1
2.2.2 Calibration ............................................................... 2-2
2.2.3 Effects of Flowrate and Gas Density ....................... 2-3
2.2.4 Measurement Results ............................................. 2-3
2.3 Electronics and Signal Processing ................................ 2-3
2.4 Temperature Control ...................................................... 2-5
3 Installation
3.1 Unpacking the Analyzer ................................................. 3-1
3.2 Mounting the Control Unit .............................................. 3-1
3.3 Electrical Connections (Rear Panel) .............................. 3-3
3.3.1 Primary Input Power .............................................. 3-4
3.3.2 Fuse Installation..................................................... 3-4
3.3.3 Analog Outputs ...................................................... 3-4
3.3.4 Alarm Relays ......................................................... 3-6
3.3.5 Digital Remote Cal Inputs ...................................... 3-7
3.3.6 Range ID Relays.................................................... 3-8
3.3.7 Network I/O............................................................ 3-8
3.3.8 RS-232 Port ........................................................... 3-9
3.3.9 Remote Probe Connector ...................................... 3-9
3.4 Testing the System ........................................................3-16
3.5 Warm Up at Power Up ...................................................3-16
4 Operation
4.1 Introduction .................................................................... 4-1
4.2 Using the Data Entry and Function Buttons................... 4-1
Teledyne Analytical Instruments
v
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 2010B
4.3 The System Function ..................................................... 4-4
4.3.1 Setting the Display ................................................. 4-5
4.3.2 Setting up an Auto-Cal........................................... 4-5
4.3.3 Password Protection .............................................. 4-6
4.3.3.1 Entering the Password................................... 4-7
4.3.3.2 Installing or Changing the Password ............. 4-7
4.3.4 Logging Out ........................................................... 4-9
4.3.5 System Self-Diagnostic Test .................................. 4-9
4.3.6 The Model Screen ................................................. 4-10
4.3.7 Checking Linearity with Algorithm.......................... 4-10
4.4 The Zero and Span Functions ....................................... 4-11
4.4.1 Zero Cal ................................................................. 4-12
4.4.1.1 Auto Mode Zeroing ........................................ 4-12
4.4.1.2 Manual Mode Zeroing.................................... 4-13
4.4.1.3 Cell Failure..................................................... 4-14
4.4.2 Span Cal ................................................................ 4-14
4.4.2.1 Auto Mode Spanning ..................................... 4-15
4.4.2.2 Manual Mode Spanning................................. 4-15
4.5 The Alarms Function...................................................... 4-16
4.6 The Range Select Function ........................................... 4-18
4.6.1 Manual (Select/Define Range) Screen .................. 4-19
4.6.2 Auto (Single Application) Screen ........................... 4-19
4.6.3 Precautions............................................................ 4-21
4.7 The Analyze Function .................................................... 4-21
4.8 Programming ................................................................. 4-22
4.8.1 The Set Range Screen .......................................... 4-23
4.8.2 The Curve Algorithm Screen ................................. 4-25
4.8.2.1 Checking the Linearization ............................ 4-25
4.8.2.2 Manual Mode Linearization............................ 4-26
4.8.2.3 Auto Mode Linearization ................................ 4-27
4.9 Special Function Setup .................................................. 4-28
4.9.1 Output Signal Reversal .......................................... 4.28
4.9.2 Special - Inverting Output ...................................... 4-28
4.9.3 Special - Polarity Coding........................................ 4-29
4.9.4 Special - Nonlinear Application Gain Preset .......... 4-29
Maintenance
5.1 Routine Maintenance ..................................................... 5-1
5.2 System Self Diagnostic Test........................................... 5-1
5.3 VFD Display ................................................................... 5-2
5.4 Fuse Replacement......................................................... 5-2
vi
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
5.5 Major Internal Components ........................................... 5-3
5.6 Cleaning......................................................................... 5-5
5.7 Phone Numbers ............................................................. 5-5
Appendix
A-1 Specifications................................................................. A-1
A-2 Recommended 2-Year Spare Parts List......................... A-3
A-3 Drawing List ................................................................... A-4
Teledyne Analytical Instruments
vii
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 2010B
DANGER
COMBUSTIBLE GAS USAGE WARNING
The customer should ensure that the principles of operating of
this equipment are well understood by the user. Misuse of this
product in any manner, tampering with its components, or
unauthorized substitution of any component may adversely
affect the safety of this instrument.
Since the use of this instrument is beyond the control of
Teledyne, no responsibility by Teledyne, its affiliates, and
agents for damage or injury from misuse or neglect of this
equipment is implied or assumed.
viii
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Introduction
1.1 Overview
TheModel2010isafamilyofsplitconfigurationconductivityanalyzers.
EachanalyzerconsistofaControlUnitsuitableforinstallationinageneral
purposearea-enclosureratedNEMA-4,andaAnalysisUnitwhichishousedin
anexplosion-proofenclosure. TheAnalysisUnitenclosureisratedforNEMA
4/7 Class I, Div. 1, Groups B,C,D and is approved by U/L and CSA.
TheAnalyticalInstrumentsModel2010BThermalConductivityAnalyzeris
aversatilemicroprocessor-basedinstrumentformeasuringacomponentgasina
backgroundgas,orinaspecificmixtureofbackgroundgases.Itcomparesthe
thermalconductivityofasamplestreamwiththatofareferencegasofknown
composition.The2010Bcan—
•
•
measure the concentration of one gas in a mixture of two gases.
measure the concentration of a gas in a specific mixture of back-
ground gases.
•
measure the purity of a sample stream containing a single impuri-
ty or a mixture of impurities.
Thestandard2010Bispre-programmedwithautomaticlinearization
algorithmsforalargenumberofgasesandgasmixtures.Thefactorycanaddto
thisdatabaseforcustomapplications,andthesophisticatedusercanaddhis
ownuniqueapplications.
ThismanualsectioncoverstheModel2010BGeneralPurposeflush-panel
andrack-mountcontrolunitsonly.
ManyoftheModel2010Bfeaturescoveredinthismanualareoptional,
selectedaccordingtothecustomersspecificapplication.Refertothespecific
modelinformationsheet(pageIV)fortheoptionsincorporatedintheinstrument.
Teledyne Analytical Instruments
Part I 1-1
Download from Www.Somanuals.com. All Manuals Search And Download.
1 Introduction
Model 2010B
1.2 Typical Applications
AfewtypicalapplicationsoftheModel2010Bare:
•
•
•
•
•
•
•
•
Power generation
Airliquefaction
Chemicalreactionmonitoring
Steel manufacturing and heat treating
Petrochemicalprocesscontrol
Qualityassurance
Refrigerationandstorage
Gasproportioningcontrol.
1.3 Main Features of the Analyzer
ThemainfeaturesoftheModel2010BThermalConductivityAnalyzer
include:
•
Three independent, user definable, analysis ranges allow up to
three different gas applications with one concentration range
each, or up to three concentration ranges for a single gas applica-
tion, or any combination.
•
Specialrecalibrationrangeformultipleapplications.Recalibrat-
ing one, recalibrates all.
•
•
Automatic, independent linearization for each range.
Auto Ranging allows analyzer to automatically select the proper
preset range for a given single application. Manual override
allows the user to lock onto a specific range of interest.
•
•
RS-232 serial digital port for use with a computer or other digital
communicationsdevice.
Six adjustable concentration set points with two alarms and a system
failurealarmrelay.
•
•
Extensive self-diagnostic testing, at startup and on demand.
A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
•
•
High resolution, accurate indication of target or impurity gas
concentration from large, bright, meter readout. (0-9999 ppm
through 0-100 % depending on types of gas involved.)
Standard, proven sensor cell design.
1-2 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
•
•
Wide range of custom applications, ranges, and linearization.
Microprocessorbasedelectronics:8-bitCMOSmicroprocessor
with 32 kB RAM and 128 kB ROM.
•
•
Autoandremotecalibrationcapabilities.
Four analog outputs: two for measurement (0–1 V dc and Isolat-
ed 4–20 mA dc) and two for range identification.
•
Compact and versatile design: Small footprint, yet internal com-
ponentsareaccessible.
1.4 Model Designations
TheModel2010Bisordinarilycustomprogrammedatthefactorytofitthe
customer’sapplication.Manyparameters,includingthenumberofchannels,the
gasapplication,thematerialsspecificationofthesamplingsystem,andothers,
areoptions.Themostcommonoptions,arecoveredinthismanual.Seethe
SpecificModelInformationchecklistinthefrontmatterofthismanualforthose
thatapplytoyourModel2010Banalyzer.Somestandardmodelsthatarenot
coveredinthismanualarelistedhere.
Models 2000A: Both analysis section and control unit are in a single
generalpurposeenclosure.
Models 2020:
Both the analysis section and control unit are in a single
explosionproofenclosure.
1.5 Operator Interface (Front Panel)
Thestandard2010Bishousedinaruggedmetalcasewithallcontrolsand
displaysaccessiblefromthefrontpanel.SeeFigure1-1.Thefrontpanelhas
thirteenbuttonsforoperatingtheanalyzer,adigitalmeter,andanalphanumeric
display.TheyaredescribedbrieflyhereandindetailintheOperationschapter
ofthismanual.
FunctionKeys:Sixtouch-sensitivemembraneswitchesareusedto
changethespecificfunctionperformedbytheanalyzer:
•
Analyze Perform analysis for target-gas content of a sample
gas.
•
System Performsystem-relatedtasks(describedindetailin
chapter 4, Operation.).
Teledyne Analytical Instruments
Part I 1-3
Download from Www.Somanuals.com. All Manuals Search And Download.
1 Introduction
Model 2010B
Outer Door
(Open)
Teledyne Analytical Instruments
Digital
M eter
Viewing
W indow
0.0
AL-1
%
Anlz
LCD
Screen
Inner Door
Latch
(Pressing the latch button
will open the inner Door)
Outer Door
Latch
Control
Panel
2010B
Therm al Conductivity Analyzer
Figure 1-1: Model 2010B Front Panel
•
•
•
•
Span
Zero
Alarms Set the alarm setpoints and attributes.
Range Set up the user definable ranges for the instrument.
Span calibrate the analyzer.
Zero calibrate the analyzer.
DataEntryKeys:Sixtouch-sensitivemembraneswitchesareusedto
inputdatatotheinstrumentviathealphanumericVFD(VacuumFluorescent
Display)display:
•
•
•
•
Left & Right Arrows
Selectbetweenfunctionscurrently
displayed on the VFD screen.
Up & Down Arrows
Increment or decrement values of
functionscurrentlydisplayed.
Enter Moves VFD on to the next screen in a series. If none
remains, returns to the Analyze screen.
Escape Moves VFD back to the previous screen in a series. If
none remains, returns to the Analyze screen.
1-4 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Digital Meter Display: ThemeterdisplayisaVFDdevicethat
produceslarge,bright,7-segmentnumbersthatarelegibleinanylighting.It
producesacontinuoustracereadoutfrom0-9999ppmoracontinuouspercent
readoutfrom1-100 %.Itisaccurateacrossallanalysisranges.
AlphanumericInterfaceScreen:TheVDFscreenisaneasy-to-use
interfacebetweenoperatorandanalyzer.Itdisplaysvalues,options,and
messagesthatgivetheoperatorimmediatefeedback.
StandbyButton:The Standbyturnsoffthedisplayandoutputs,
butcircuitryisstilloperating.
CAUTION: The power cable must be unplugged to fully
disconnect power from the instrument. When
chassis is exposed or when access door is open
and power cable is connected, use extra care to
avoid contact with live electrical circuits.
AccessDoor:Foraccesstothethermalconductivitysensororthefront
panelelectronics,thefrontpanelswingsopenwhenthelatchintheupperright
cornerofthepanelispressedallthewayinwithanarrowgaugetool.Accessing
themainelectronicscircuitboardrequiresunfasteningrearpanelscrewsand
slidingtheelectronicsdraweroutofthecase.(Seechapter5.)
1.6 Recognizing Difference Between LCD &
VFD
LCD(LiquidCrystalDisplay)hasGREENbackgroundwithBLACK
characters. VFD has DARK background withGREEN characters. In the case
ofVFD (VacuumFluorescentDisplay)-NOCONTRASTADJUSTMENTIS
NEEDED.
1.7 Equipment Interface (Rear Panel)
Therearpanel,showninFigure1-2,containstheelectricalconnectorsfor
externalinputandoutput.Theconnectorsaredescribedbrieflyhereandindetail
inchapter3,Installation.
Teledyne Analytical Instruments
Part I 1-5
Download from Www.Somanuals.com. All Manuals Search And Download.
1 Introduction
Model 2010B
Figure 1-2: Model 2010B Rear Panel
•
•
Power Connection
Analog Outputs
85-250 V AC power source.
0-1 V dc concentration plus 0-1 V dc
range ID, and isolated 4-20 mA dc plus
4-20 mA dc range ID.
1-6 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
•
•
•
•
•
Alarm Connections
2 concentration alarms and 1 system
alarm.
RS-232 Port
Serialdigitalconcentrationsignaloutput
and control input.
Remote Probe
Remote Span/Zero
Used in the 2010B to interface the
external Analysis Unit.
Digital inputs allow external control of
analyzercalibration.
Calibration Contact To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
•
•
Range ID Contacts
Network I/O
Four separate, dedicated,
range-identificationrelaycontacts(01,
02, 03,CAL).
Serialdigitalcommunicationsforlocal
network access. For future expansion.
Not implemented at this printing.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 2010B is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
Teledyne Analytical Instruments
Part I 1-7
Download from Www.Somanuals.com. All Manuals Search And Download.
1 Introduction
Model 2010B
1-8 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Operational Theory
2.1 Introduction
The analyzer is composed of two subsystems:
1. ThermalConductivitySensor
2. Electronic Signal Processing, Display and Control.
The sensor is a thermal conductivity comparator that continuously
compares the thermal conductivity of the sample gas with that of a reference
gas having a known conductivity.
The electronic signal processing, display and control subsystem simpli-
fies operation of the analyzer and accurately processes the sampled data. A
microprocessorcontrolsallsignalprocessing, input/output, anddisplay
functions for the analyzer.
2.2 Sensor Theory
For greater clarity, Figure 2-1 presents two different illustrations, (a)
and (b), of the operating principle of the thermal conductivity cell.
2.2.1 Sensor Configuration
The thermal conductivity sensor contains two chambers, one for the
reference gas of known conductivity and one for the sample gas. Each
chamber contains a pair of heated filaments. Depending on its thermal
conductivity, each of the gases conducts a quantity of heat away from the
filaments in its chamber. See Figure 2-1(a).
The resistance of the filaments depends on their temperature. These
filaments are parts of the two legs of a Wheatstone bridge circuit that unbal-
ances if the resistances of its two legs do not match. See Figure 2-1(b).
Teledyne Analytical Instruments
Part I 2-1
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Operational Theory
Model 2010B
Figure 2-1: Thermal Conductivity Cell Operating Principle
If the thermal conductivities of the gases in the two chambers are
different, the Wheatstone bridge circuit unbalances, causing a current to flow
in its detector circuit. The amount of this current can be an indication of the
amount of impurity in the sample gas, or even an indication of the type of
gas, depending on the known properties of the reference and sample gases.
The temperature of the measuring cell is regulated to within 0.1 °C by a
sophisticated control circuit. Temperature control is precise enough to com-
pensate for diurnal effects in the output over the operating ranges of the
analyzer. (See Specifications in the Appendix for details.)
2.2.2 Calibration
Because analysis by thermal conductivity is not an absolute measure-
ment, calibration gases of known composition are required to fix the upper
and lower parameters (“zero” and “span”) of the range, or ranges, of analy-
sis. These gases must be used periodically, to check the accuracy of the
analyzer.
During calibration, the bridge circuit is balanced, with zero gas against
the reference gas, at one end of the measurement range; and it is sensitized
with span gas against the reference gas at the other end of the measurement
range. The resulting electrical signals are processed by the analyzer electron-
ics to produce a standard 0-1V, or an isolated 4–20 mA dc, output signal, as
described in the next section.
Teledyne Analytical Instruments
2-2 Part I
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
2.2.3 Effects of Flowrate and Gas Density
Because the flowrate of the gases in the chambers affects their cooling
of the heated filaments, the flowrate in the chambers must be kept as equal,
constant, and low as possible.
When setting the sample and reference flowrate, note that gases lighter
than air will have an actual flowrate higher than indicated on the flowmeter,
while gases heavier than air will have an actual flowrate lower than indi-
cated. Due to the wide range of gases that are measured with the Thermal
Conductivity Analyzer, the densities of the gases being handled may vary
considerably.
Then, there are limited applications where the reference gas is in a
sealed chamber and does not flow at all. These effects must be taken in
consideration by the user when setting up an analysis.
2.2.4 Measurement Results
Thermal conductivity measurements are nonspecific by nature. This fact
imposes certain limitations and requirements. If the user intends to employ
the analyzer to detect a specific component in a sample stream, the sample
must be composed of the component of interest and one other gas (or spe-
cific, and constant, mixture of gases) in order for the measured heat-transfer
differences to be nonambiguous.
If, on the other hand, the user is primarily interested in the purity of a
process stream, and does not require specific identification of the impurity,
the analyzer can be used on more complex mixtures.
2.3 Electronics and Signal Processing
The Model 2010B Thermal Conductivity Analyzer uses an 8031
microcontroller, Central Processing Unit—(CPU) with 32 kB of RAM and
128 kB of ROM to control all signal processing, input/output, and display
functions for the analyzer. System power is supplied from a universal power
supply module designed to be compatible with any international power
source. (See Major Internal Components in chapter 5 Maintenance for the
location of the power supply and the main electronic PC boards.)
The signal processing electronics including the microprocessor, analog
to digital, and digital to analog converters are located on the Motherboard at
the bottom of the case. The Preamplifier board is mounted on top of the
Motherboard as shown in the figure 5.4. These boards are accessible after
Teledyne Analytical Instruments
Part I 2-3
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Operational Theory
Model 2010B
removing the back panel. Figure 2-2 is a block diagram of the Analyzer
electronics.
Variable
Differential
Amplifier
Gain
Amplifier
Thermistor
A to D
Converter
To CPU
M
U
X
Sensor
Heater
Auto-
Range
0-1 V dc
Concentration
and Range
Temperature
Control
Digitial to
Analog
Converter
(DAC)
Coarse
Adjustment
4-20 mA dc
Concentration
and Range
Fine
Adjustment
Temperature
Control
Heater
Alarm 1
Alarm 2
Selt-Test Signal to MUX
System
Failure
Alarm
RS-232
Keyboard
Central
Processing
Unit
Range
Contacts (4)
(CPU)
Displays
Processing
External
Valve
Control
Remote Span
Control
Power
Supply
Remote Zero
Control
A to D Conv
Cal
Contact
Figure 2-2: Block Diagram of the Model 2010B Electronics
Teledyne Analytical Instruments
2-4 Part I
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
In the presence of dissimilar gases the sensor generates a differential
voltageacrossitsoutputterminals. Adifferentialamplifierconvertsthis
signal to a unipolar signal, which is amplified in the second stage, variable
gain amplifier, which provides automatic range switching under control of
the CPU. The output from the variable gain amplifier is sent to an 18 bit
analog to digital converter.
The digital concentration signal along with input from the Gas Selector
Panel is processed by the CPU and passed on to the 12-bit DAC, which
outputs 0-1 V dc Concentration and Range ID signals. An voltage-to-current
converter provides 4-20 mA dc concentration signal and range ID outputs.
The CPU also provides appropriate control signals to the Displays,
Alarms, and External Valve Controls, and accepts digital inputs for external
Remote Zero and Remote Span commands. It monitors the power supply
through an analog to digital converter as part of the data for the system
failurealarm.
The RS-232 port provides two-way serial digital communications to
and from the CPU. These, and all of the above electrical interface signals are
described in detail in chapter 3 Installation.
2.4. Temperature Control
For accurate analysis the sensor of this instrument is temperature con-
trolled to 50oC.
The Temperature Control keeps the temperature of the measuring cell
regulated to within 0.1 degree C. A thermistor is used to measure the tem-
perature, and a zero-crossing switch regulates the power in a cartridge-type
heater. The result is a sensor output signal that is temperature independent.
A second temperature control system is used to maintain the Analysis
Unit internal temperature at 220C minimum.
Teledyne Analytical Instruments
Part I 2-5
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Operational Theory
Model 2010B
Teledyne Analytical Instruments
2-6 Part I
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Installation
Installation of the Model 2010B Analyzer includes:
1. Unpacking
2. Mounting
3. Gas connections
4. Electricalconnections
5. Testing the system.
3.1 Unpacking the Analyzer
The analyzer is shipped ready to install and prepared for operation.
Carefully unpack the analyzer and inspect it for damage. Immediately report
any damage to the shipping agent.
3.2 Mounting the Control Unit
The Model 2010B Control Unit is designed for bulkhead mounting in
general purpose area, NOT for hazardous environments of any type. The
Control Unit is for indoor/outdoor use.
Figure 3-1 is an illustration of the 2010B standard front panel and
mounting bezel. There are four mounting holes—one in each corner of the
rigid frame. See the outline drawing, at the back of this manual for overall
dimensions.
All operator controls are mounted on the inner control panel, which is
hinged on the left edge and doubles as a door to provide access to the internal
components of the instrument. The door will swing open when the button of
the latch is pressed all the way in with a narrow gauge tool (less than 0.18
inch wide), such as a small hex wrench or screwdriver Allow clearance for
the door to open in a 90-degree arc of radius 11.75 inches. See Figure 3-2.
Teledyne Analytical Instruments
Part I 3-1
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Installation
Model 2010B
NPT Fittings
supplied by
customer
Viewing
W indow
0.0
AL-1
% Anlz
Outer Door
Latch
HHinge
Figure 3-1: Front Panel of the Model 2010B Control Unit
Figure 3-2: Required Front Door Clearance
H
3-2 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
3.3 Electrical Connections (Rear Panel)
Figure 3-3 shows the Model 2010B Electrical Connector Panel. There
are terminal blocks for connecting power, communications, and both digital
and analog concentration outputs.
For safe connections, ensure that no uninsulated wire extends outside of
the connectors they are attached to. Stripped wire ends must insert com-
pletely into terminal blocks. No uninsulated wiring should be able to come in
contact with fingers, tools or clothing during normal operation.
Figure 3-3: Rear Panel of the Model 2010B
3.3.1 Primary Input Power
The power terminal block and fuse receptacles are located in the same
rearpanelassembly.
Teledyne Analytical Instruments
Part I 3-3
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Installation
Model 2010B
DANGER: POWER IS APPLIED TO THE INSTRUMENT'S CIR-
CUITRY AS LONG AS THE INSTRUMENT IS CON-
NECTED TO THE POWER SOURCE. THE RED I/O
SWITCH ON THE FRONT PANEL IS FOR SWITCH-
ING POWER ON OR OFF TO THE DISPLAYS AND
OUTPUTS ONLY.
NOTE:
AC POWER MAY BE PRESENT ON THE RELAY
CONTACTS WHEN THE POWER CORD IS RE-
MOVED!
The Control Unit is universal power 100-240V, 50-60 Hz. The Analy-
sis Unit requires 110 or 220 VAC and is selectable via switch located inside
theexplosion-proofenclosure.
3.3.2 Fuse Installation
The fuse receptacles accepts European size fuses only (5x20mm). Both
sides of the line are fused. If the European size fuses are selected, both sides
at the line will be fused. Be sure to install the proper fuse as part of installa-
tion. (See Fuse Replacement in chapter 5, maintenance.)
3.3.3 Analog Outputs
There are four DC output signal connectors with spring terminals on the
panel. There are two wires per output with the polarity noted. See Figure 3-
4. The outputs are:
0–1 V dc % of Range: Voltageriseslinearlywithincreasingconcentration,
from 0 V at 0 concentration to 1 V at full scale.
(Full scale = 100% of programmable range.)
0–1 V dc Range ID:
0.25 V = Range 1, 0.5 V = Range 2, 0.75 V =
Range 3, 1 V = Cal Range.
4–20 mA dc % Range: Current rises linearly with concentration, from 4
mA at 0 concentration to 20 mA at full scale. (Full
scale = 100% of programmable range.)
4–20 mA dc Range ID: 8 mA = Range 1, 12 mA = Range 2, 16 mA =
Range 3, 20 mA = Range 4.
3-4 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Figure 3-4: Analog Output Connections
Examples:
The analog output signal has a voltage which depends on gas concen-
tration relative to the full scale of the range. To relate the signal output to the
actual concentration, it is necessary to know what range the instrument is
currently on, especially when the analyzer is in the autoranging mode.
The signal output for concentration is linear over the currently selected
analysis range. For example, if the analyzer is set on a range that was defined
as 0–10 % hydrogen, then the output would be as shown in Table 3-1.
Table 3-1: Analog Concentration Output—Example
Percent
Voltage Signal Current Signal
Hydrogen Output (V dc)
Output (mA dc)
0
1
2
3
4
5
6
7
8
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
4.0
5.6
7.2
8.8
10.4
12.0
13.6
15.2
16.8
18.4
20.0
9
10
(Continued)
Teledyne Analytical Instruments
Part I 3-5
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Installation
Model 2010B
To provide an indication of the range, the Range ID analog output
terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 3-2 gives the
range ID output for each analysis range.
Table 3-2: Analog Range ID Output—Example
Range
Voltage (V)
Current (mA) Application
Range 1
0.25
8
12
16
20
0-1 % H2 in N2
0-10 % H2 in N2
0-1 % H2 in Air
0-1 % H2 in N2
Range 2
0.50
0.75
1.00
Range 3
Range 4 (Cal)
3.3.4 Alarm Relays
The three alarm-circuit connectors are spring terminals for making
connections to internal alarm relay contacts. Each provides a set of Form C
contacts for each type of alarm. Each has both normally open and normally
closed contact connections. The contact connections are indicated by dia-
grams on the rear panel. They are capable of switching up to 3 amperes at
250 V ac into a resistive load. See Figure 3-5. The connectors are:
Threshold Alarm 1: • Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentrationisbelowthreshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
Threshold Alarm 2: • Can be configured as high (actuates when concen-
tration is above threshold), or low (actuates when
concentrationisbelowthreshold).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).
SystemAlarm:
Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently
configured as fail-safe and latching. Cannot be de-
feated.
Actuates when cell can not balance during zero
calibration.
Actuates when span parameter out off its limited
parameter.
3-6 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Actuates when self test fails.
(Reset by pressing I/O button to remove power. Then
press I/O again and any other button EXCEPT
Systemtoresume.
Further detail can be found in chapter 4, section 4-5.
Figure 3-5: Types of Relay Contacts
3.3.5 Digital Remote Cal Inputs
Accept 0 V (off) or 24 V dc (on) inputs for remote control of calibra-
tion. (See Remote Calibration Protocol below.)
Zero:
Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the Zero mode. Either side may be
grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appropri-
ately. See 3.3.9 Remote Probe Connector. (With the –C
option,theinternalvalvesoperateautomatically.)
Span:
Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the Span mode. Either side may be
grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appropri-
ately. See 3.3.9 Remote Probe Connector. (With the –C
option,theinternalvalvesoperateautomatically.)
Cal Contact: This relay contact is closed while analyzer is spanning
and/or zeroing. (See Remote Calibration Protocol below.)
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 2010B Analyzer, the customer's controller must
monitor the Cal Relay Contact.
Teledyne Analytical Instruments
Part I 3-7
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Installation
Model 2010B
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It
will ignore your request to calibrate, and it will not remember that request.
Once a zero or span command is sent, and acknowledged (contact
closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
closeagain.
When the contact is closed, the display would display the last reading of
the gas concentration value and output signal would output the last reading
from the sample gas (SAMPLE and HOLD).
Forexample:
1) Test the CRC. When the CRC is open, Send a zero command
until the CRC closes (The CRC will close quickly.)
2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC
closes. (The CRC will close quickly.)
4) When the CRC closes, remove the span command.
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The Remote Probe connector (paragraph 3.3.9) provides
signals to operate the zero and span gas valves synchro-
nously. However, if you have the –C, -P, or -Q Internal valve
option, which includes zero and span gas inputs, the 2010B
automatically selects the zero, span and sample gas flow.
3.3.6 Range ID Relays
Four dedicated Range ID relay contacts. For any single application they
are assigned to relays in ascending order. For example: if all ranges have the
same application, then the lowest range is assigned to the Range 1 ID relay,
and the highest range is assigned to the Range 3 ID relay. Range 4 is the Cal
Range ID relay.
3.3.7 NetworkI/O
A serial digital input/output for local network protocol. At this printing,
this port is not yet functional. It is to be used in future versions of the instru-
ment.
3-8 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
3.3.8 RS-232 Port
Part I: Control Unit
The digital signal output is a standard RS-232 serial communications
port used to connect the analyzer to a computer, terminal, or other digital
device. It requires a standard 9-pin D connector.
Output: The data output is status information, in digital form, updated
every two seconds. Status is reported in the following order:
•
•
The concentration in ppm or percent
The range in use (01 = Range 1, 02 = Range 2, 03 = Range 3,
CAL = Range 4)
•
•
•
The scale of the range (0-100 %, etc)
Which alarms—if any—are disabled (AL–x DISABLED)
Which alarms—if any—are tripped (AL–x ON).
Each status output is followed by a carriage return and line feed.
Input: The input functions using RS-232 that have been implemented
to date are described in Table 3-3.
Table 3-3: Commands via RS-232 Input
Command
as<enter>
az<enter>
rp<enter>
Description
Immediatelystartsanautospan.
Immediatelystartsanautozero.
Allows reprogramming of the APPLICATION (gas use)
and ALGORITHM (linearization) System functions.
st<enter>
Toggling input. Stops/Starts any status message output from
the RS-232, until st<enter> is sent again.
rm1<enter>
rm2<enter>
rm3<enter>
rm4<enter>
ra<enter>
Range manual 1
Range manual 2
Range manual 3
Range manual CAL
Range auto
Implementation: The RS-232 protocol allows some flexibility in its
implementation. Table 3-4 lists certain RS-232 values that are required by
the Model 2010B implementation.
Teledyne Analytical Instruments
Part I 3-9
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Installation
Model 2010B
Table 3-4: Required RS-232 Options
Parameter
Setting
Baud 2400
Byte
Parity
Stop Bits
8 bits
none
1
MessageInterval
2 seconds
3.3.9 Remote Probe Connector
The 2010B is a single-split configuration analyzer, the Remote Probe
connector is used to interface the Analysis Unit. See Figure 3-6.
SPA N IN
SAM PLE IN
ZE RO IN
EXH AUST
+15V
Turn cw to hold
ccw to
loosen w ire.
Switching
G round
Insert wire
here.
Figure 3-6: Remote Probe Connector Pinouts
The voltage from these outputs is nominally 0 V for the OFF and
15 V dc for the ON conditions. The maximum combined current that can be
pulled from these output lines is 100 mA. (If two lines are ON at the same
time, each must be limited to 50 mA, etc.) If more current and/or a different
voltage is required, use a relay, power amplifier, or other matching circuitry
to provide the actual driving current.
In addition, each individual line has a series FET with a nominal ON
resistance of 5 ohms (9 ohms worst case). This could limit the obtainable
voltage, depending on the load impedance applied. See Figure 3-7.
3-10 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Figure 3-7: FET Series Resistance
3.4 Testing the System
Before plugging the instrument into the power source:
•
•
•
Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.
Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors
Check that the pressure and flow of all gases are within the
recommended levels, and appropriate for your application.
Power up the system, and test it by performing the following
operations:
1. Repeat the Self-Diagnostic Test as described in chapter 4, section
4.3.5.
3.5 Warm Up at Power Up
Every time the unit is turned on, the instrument stays with the introduc-
tion screen for thirty minutes. This is to allow the cell to come up to tem-
perature (50oC). The only way to bypass this warm up period is by pressing
any key once, such as the Enter key.
The instrument warms up for half an hour so that it will not receive a
remote calibration signal, send false readings to a monitor system, or, again,
be calibrated by an untrained operator while the cell is cold.
NOTE:There is no feedback on whether the working temperature
has been achieved by cell to the software. If instrument power
is interrupted for only a brief time, the instrument will wait
thirty minutes again. Press the ENTER key to bypass self-
diagnostic mode.
Teledyne Analytical Instruments
Part I 3-11
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Installation
Model 2010B
3-12 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Operation
4.1 Introduction
AlthoughtheModel2010Bisusuallyprogrammedtoyourapplicationat
thefactory,itcanbefurtherconfiguredattheoperatorlevel,oreven,cautious-
ly,reprogrammed.Dependingonthespecificsoftheapplication,thismight
includeallorasubsetofthefollowingprocedures:
•
Settingsystemparameters:
•
Establish a security password, if desired, requiring Operator
to log in.
•
Establish and start an automatic calibration cycle, if desired.
•
RoutineOperation:
•
•
•
Calibratetheinstrument.
Choose autoranging or select a fixed range of analysis.
Set alarm setpoints, and modes of alarm operation (latching,
fail-safe,etc).
•
Program/Reprogramtheanalyzer:
•
•
Define new applications.
Linearize your ranges.
Ifyouchoosenottousepasswordprotection,thedefaultpasswordis
automaticallydisplayedonthepasswordscreenwhenyoustartup,andyou
simplypressEnterforaccesstoallfunctionsoftheanalyzer.
4.2 Using the Data Entry and Function
Buttons
DataEntryButtons:The<>buttonsselectoptionsfromthemenu
currentlybeingdisplayedontheVFDscreen.Theselectedoptionblinks.
Whentheselectedoptionincludesamodifiableitem,the∆∇arrowbuttons
canbeusedtoincrementordecrementthatmodifiableitem.
Teledyne Analytical Instruments
Part I 4-1
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
The EnterbuttonisusedtoacceptanynewentriesontheVFDscreen.
The EscapebuttonisusedtoabortanynewentriesontheVFDscreenthatare
not yet accepted by use of the Enterbutton.
Figure4-1showsthehierarchyoffunctionsavailabletotheoperatorviathe
functionbuttons.Thesixfunctionbuttonsontheanalyzerare:
• Analyze. This is the normal operating mode. The analyzer
monitors the thermal conductivity of the sample, displays the
percent or parts-per-million of target gas or contamination, and
warns of any alarm conditions.
• System. The system function consists of nine subfunctions.
Four of these are for ordinary setup and operation:
•
•
•
•
Setup an Auto-Cal
Assign Passwords
Log out to secure system
InitiateaSelf-Test
Three of the subfunctions do auxiliary tasks:
•
•
Checking model and software version
Displaymoresubfunctions
Two of these are for programming/reprogramming the analyzer:
•
Define gas applications and ranges (Refer to programming
section, or contact factory.)
•
Use the Curve Algorithm to linearize output. (Refer to
programmingsection, orcontactfactory.)
• Zero. Used to set up a zero calibration.
• Span. Used to set up a span calibration.
• Alarms. Used to set the alarm setpoints and determine whether
each alarm will be active or defeated, HI or LO acting, latching,
and/orfail-safe.
• Range. Used to set up three analysis ranges that can be
switched automatically with autoranging or used as individual
fixedranges.
Anyfunctioncanbeselectedatanytimebypressingtheappropriatebutton
(unlesspasswordrestrictionsapply).Theorderaspresentedinthismanualis
appropriateforaninitialsetup.
Eachofthesefunctionsisdescribedingreaterdetailinthefollowingproce-
dures.TheVFDscreentextthataccompanieseachoperationisreproduced,at
theappropriatepointintheprocedure,inaMonospacedtypestyle.Push-
buttonnamesareprintedinObliquetype.
4-2 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
System
Set LCD
CONTRAST
Contrast Function is DISABLED
Contrast
(Refer to Section 1.6)
Span/Zero
Off/On
Span/Zero
Timing
Span/Zero
Off/On
AUTO-CAL
PASSWORD
LOGOUT
Enter
Password
Change Yes
Yes/No
Change
Password
Verify
Password
Secure Sys &
Analyze Only
MORE
Show Model
and Version
MODEL
Select
Range
Define
Appl/Range
APPLICATION
SELF-TEST
Self-Test in
Progress
Slef-Test
Results
Verify
Points
Ver
Enter
Input/Output
Values
Gas Use
Range
Select
Verify/Setup
Select
Range
Man
ALGORITHM
Enter
Enter
Auto/Manual
Linearity Cal
Set
Select Linrty
Span Values
Auto
Zero in
Progress
Auto/Manual
Zero Select
Zero
Auto/Manual
Span Select
Span Value
Set
Span in
Progress
Span
Select
Range
Gas Use
Range
% / ppm
Select
Setpoints &
Attributes
Alarms
Man
Define
Range
Select
Range
Auto/Manual
Range Adj
Range
Gas
Application
Auto
Analyze
Sample
Analyze
Figure 4-1: Hierarchy of Functions and Subfunctions
Teledyne Analytical Instruments
Part I 4-3
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
4.3 The System Function
ThesubfunctionsoftheSystemfunctionaredescribedbelow.Specific
proceduresfortheirusefollowthedescriptions:
•
•
AUTO-CAL: Used to define an automatic calibration sequence
and/or start an AUTO-CAL.
PWD: Security can be established by choosing a 3 digit
password (PWD) from the standard ASCII character set. Once a
unique password is assigned and activated, the operator MUST
enter the UNIQUE password to gain access to setup functions
which alter the instrument's operation.
•
•
•
•
LOGOUT: Logging out prevents an unauthorized tampering
withanalyzersettings.
MORE: Select and enter MORE to get a new screen with
additionalsubfunctionslisted.
MODEL: Displays Manufacturer, Model, and Software Version
ofinstrument.
APPLICATION: A restricted function, not generally accessed by
theenduser.Usedtodefineuptothreeanalysisrangesanda
calibrationrange(includingimpurity,background,lowendofrange,
highendofrange,and%orppmunits).
•
•
SELF-TEST: The instrument performs a self-diagnostic test to
check the integrity of the power supply, output boards, sensor
cell,andpreamplifiers.
ALGORITHM: A restricted function, not generally accessed by the
end user. Used to linearize the output for the range of interest.
4-4 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
4.3.1 Setting the Display
ContrastFunctionisDISABLED
(Refer to Section 1.6)
IfyoucannotreadanythingontheVFDafterfirstpoweringup:
1. ObserveLEDreadout.
a. If LED meter reads 8.8.8.8.8., go to step 3.
b. If LED meter displays anything else, go to step 2.
2. Press I/O button twice to turn Analyzer OFF and ON again. LED
meter should now read 8.8.8.8.8.. Go to step 3.
4.3.2 Setting up an AUTO-CAL
Whenproperautomaticvalvingisconnected(seechapter3,installation),
theAnalyzercancycleitselfthroughasequenceofstepsthatautomaticallyzero
andspantheinstrument.
Note: Before setting up an AUTO-CAL, be sure you understand the
Zero and Span functions as described in section 4.4, and
follow the precautions given there.
Note: If you require highly accurate AUTO-CAL timing, use external
AUTO-CAL control where possible. The internal clock in the
Model 2010B is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
Note: If all your ranges are for the same gas application, then AUTO-
CAL will calibrate whichever range you are in at the scheduled
time for automatic calibration.
Note: If your ranges are configured for different applications, then
AUTO-CAL will calibrate all of the ranges simultaneously (by
calibrating the Cal Range).
To setup an AutoCal cycle:
ChooseSystemfromtheFunctionbuttons.TheVFD willdisplayfive
subfunctions.
CONTRAST AUTOCAL
PWD LOGOUT MORE
Contrast Function is DISABLED
(Refer to Section 1.6)
Teledyne Analytical Instruments
Part I 4-5
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
Use < > arrows to blink AUTOCAL, and press Enter. A new screen for
ZERO/SPAN set appears.
ZERO in Ød Øh off
SPANin Ød Øhoff
Press < > arrows to blink ZERO(or SPAN), then press Enter again.
(You won’t be able to set OFF to ON if a zero interval is entered.) A Span
Every ... (or Zero Every ...) screen appears.
Zeroschedule:OFF
Day: ØdHour: Øh
Use ∆∇arrows to set an interval value, then use < >arrowstomovetothe
start-timevalue.Use∆∇arrowstosetastart-timevalue.
To turn ON the SPAN and/or ZERO cycles (to activate AUTOCAL):
Press System again, choose AUTOCAL, and press Enter again. When the
ZERO/SPAN values screen appears, use the < > arrows to blink the ZERO
(or SPAN) and press Enter to go to the next screen. Use < > to select OFF/
ONfield. Use ∆∇ arrows to set the OFF/ONfield to ON. You can now turn
thesefieldsONbecausethereisanonzerospanintervaldefined.
4.3.3 Password Protection
Beforeauniquepasswordisassigned,thesystemassignsTAIbydefault.
Thispasswordwillbedisplayedautomatically.Theoperatorjustpressesthe
Enterkeytobeallowedtotalaccesstotheinstrument’sfeatures.
Ifapasswordisassigned,thensettingthefollowingsystemparameterscan
bedoneonlyafterthepasswordisentered:alarmsetpoints,assigninganew
password,range/applicationselections,andcurvealgorithmlinearization.
(APPLICATIONandALGORITHMarecoveredintheprogrammingsection.)
However,theinstrumentcanstillbeusedforanalysisorforinitiatingaself-test
withoutenteringthepassword.Todefeatsecuritythepasswordmustbe
changed back to TAI.
NOTE: If you use password security, it is advisable to keep a copy of
the password in a separate, safe location.
4-6 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
4.3.3.1
EnteringthePassword
Toinstallanewpasswordorchangeapreviouslyinstalledpassword,you
mustkeyinandENTERtheoldpasswordfirst.Ifthedefaultpasswordisin
effect, pressingtheENTERbuttonwillenterthedefaultTAIpasswordforyou.
Press System to enter the System mode.
ContrastFunctionisDISABLED
CONTRAST AUTOCAL
PWD LOGOUT MORE
(Refer to Section 1.6)
Use the < > arrow keys to scroll the blinking over to PWD, and press
Enterto select the password function. Either the default TAI password or AAA
placeholdersforanexistingpasswordwillappearonscreendependingon
whetherornotapasswordhasbeenpreviouslyinstalled.
Enterpassword:
TAI
or
Enterpassword:
AAA
Thescreenpromptsyoutoenterthecurrentpassword.Ifyouarenotusing
password protection, press Enter to accept TAI as the default password. If a
passwordhasbeenpreviouslyinstalled,enterthepasswordusingthe<>arrow
keystoscrollbackandforthbetweenletters, andthe∆∇arrowkeystochange
the letters to the proper password. Press Enter to enter the password.
Inafewseconds,youwillbegiventheopportunitytochangethispass-
word or keep it and go on.
ChangePassword?
<ENT>=Yes <ESC>=No
Press Escape to move on, or proceed as in Changing the Password,
below.
4.3.3.2
InstallingorChangingthePassword
Ifyouwanttoinstallapassword,orchangeanexistingpassword,proceed
asaboveinEnteringthePassword. Whenyouaregiventheopportunityto
changethepassword:
ChangePassword?
<ENT>=Yes <ESC>=No
PressEntertochange the password(eitherthe defaultTAI or the previ-
ouslyassignedpassword),orpressEscapetokeeptheexistingpasswordand
moveon.
Teledyne Analytical Instruments
Part I 4-7
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
IfyouchoseEntertochangethepassword,thepasswordassignment
screenappears.
Selectnewpassword
TAI
Enter the password using the < > arrowkeys tomove backand forth
betweentheexistingpasswordletters,andthe∆∇arrowkeystochangethe
letterstothenewpassword.Thefullsetof94charactersavailableforpassword
useareshowninthetablebelow.
Characters Available for Password Definition:
A
K
U
_
i
s
}
)
3
=
B
L
V
`
j
t
→
*
4
>
C
M
W
a
k
u
!
+
5
?
D
N
X
b
l
v
"
'
6
@
E
O
Y
c
m
w
#
-
7
F
P
Z
d
n
x
$
.
8
G
Q
[
e
o
y
%
/
9
H
R
¥
f
p
z
&
0
:
I
S
]
g
q
{
'
1
;
J
T
^
h
r
|
(
2
<
Whenyouhavefinishedtypingthenewpassword,pressEnter.Averifica-
tionscreenappears.Thescreenwillpromptyoutoretypeyourpasswordfor
verification.
EnterPWDToVerify:
AAA
Use the arrow keys to retype your password and press Enter when
finished.Yourpasswordwillbestoredinthemicroprocessorandthesystemwill
immediatelyswitchtotheAnalyzescreen,andyounowhaveaccesstoall
instrumentfunctions.
Ifallalarmsaredefeated, theAnalyzescreenappearsas:
1.95 %H2inN2
nR1: Ø1ØAnlz
Ifanalarmistripped,thesecondlinewillchangetoshowwhichalarmitis:
1.95 %H2inN2
AL1
4-8 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
NOTE:If you log off the system using the LOGOUT function in the
system menu, you will now be required to re-enter the pass-
word to gain access to Alarm, and Range functions.
4.3.4 Logging Out
TheLOGOUTfunction providesaconvenientmeansofleavingtheanalyzer
inapasswordprotectedmodewithouthavingtoshuttheinstrumentoff.By
enteringLOGOUT,youeffectivelylogofftheinstrumentleavingthesystem
protectedagainstuseuntilthepasswordisreentered.Tologout,pressthe
SystembuttontoentertheSystemfunction.
CONTRAST AUTOCAL
PWD LOGOUT MORE
ContrastFunctionisDISABLED
(Refer to Section 1.6)
Use the < > arrow keys to position the blinking over the LOGOUT func-
tion,andpressEntertoLogout.Thescreenwilldisplaythemessage:
Protecteduntil
passwordentered
4.3.5 System Self-Diagnostic Test
TheModel2010Bhasabuilt-inself-diagnostictestingroutine.Pre-pro-
grammedsignalsaresentthroughthepowersupply,outputboard,preamp
boardandsensorcircuit.Thereturnsignalisanalyzed,andattheendofthetest
thestatusofeachfunctionisdisplayedonthescreen, eitherasOK orasa
number between 1 and 1024. (See System Self Diagnostic Test in chapter 5
fornumbercode.)Ifanyofthefunctionsfails,theSystemAlarmistripped.
Note: The sensor will always show failed unless identical gas is
present in both channels at the time of the SELF-TEST.
Theselfdiagnosticsarerunautomaticallybytheanalyzerwheneverthe
instrumentisturnedon,butthetestcanalsoberunbytheoperatoratwill.To
initiateaselfdiagnostictestduringoperation:
PresstheSystembuttontostarttheSystemfunction.
CONTRAST AUTOCAL
PWD LOGOUT MORE
Contrast Function is DISABLED
(Refer to Section 1.6)
Use the < > arrow keys to blink MORE, then press Enter.
MODEL APPLICATION
SELFTEST ALGORITHM
Use the < > arrow keys again to move the blinking to the SELFTEST
andpressEnter.Thescreenwillfollowtherunningofthediagnostic.
Teledyne Analytical Instruments
Part I 4-9
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
RUNNING DIAGNOSTIC
TestingPreampCell
Whenthetestingiscomplete,theresultsaredisplayed.
Power:OK Analog:OK
Cell: 2 Preamp:3
ThemoduleisfunctioningproperlyifitisfollowedbyOK.Anumber
indicatesaprobleminaspecificareaoftheinstrument.RefertoChapter5
MaintenanceandTroubleshootingfornumber-codeinformation.Theresults
screenalternatesforatimewith:
PressAnyKey
ToContinue...
ThentheanalyzerreturnstotheinitialSystemscreen.
4.3.6 The Model Screen
Move the < > arrow key to MORE and press Enter. With MODEL
blinking,pressEnter.Thescreendisplaysthemanufacturer,model,andsoft-
wareversioninformation.
4.3.7 Checking Linearity with ALGORITHM
From the System Function screen, select ALGORITHM, and press
Enter.
selrngtosetalgo:
> Ø1 Ø2 Ø3 <
Use the < > keys to select the range: 01, 02, or 03. Then press Enter.
Gas Use: H2 N2
Range:
Ø
10%
PressEnteragain.
Algorithmsetup:
VERIFY SETUP
Select and EnterVERIFY to check whether the linearization has been
accomplishedsatisfactorily.
Dpt INPUT OUTPUT
Ø Ø.ØØ Ø.ØØ
4-10 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Theleftmostdigit(underDpt)isthenumberofthedatapointbeingmoni-
tored.Usethe∆∇keystoselectthesuccessivepoints.
TheINPUTvalueistheinputtothelinearizer.Itisthesimulatedoutputof
theanalyzer.Youdonotneedtoactuallyflowgas.
The OUTPUT value is the output of the linearizer. It should be the ACTU-
ALconcentrationofthespangasbeingsimulated.
IftheOUTPUTvalueshownisnotcorrect,thelinearizationmustbe
corrected. PressESCAPEtoreturntothepreviousscreen. SelectandEnter
SET UP to Calibration Mode screen.
Selectalgorithm
mode : AUTO
There are two ways to linearize: AUTO and MANUAL: The auto mode
requiresasmanycalibrationgasesastherewillbecorrectionpointsalongthe
curve.Theuserdecidesonthenumberofpoints,basedontheprecisionre-
quired.
Themanualmodeonlyrequiresenteringthevaluesforeachcorrection
pointintothemicroprocessorviathefrontpanelbuttons.Again,thenumberof
pointsrequiredisdeterminedbytheuser.
4.4 The Zero and Span Functions
(1)TheModel2010Bcanhaveasmanyasthreeanalysisrangesplusa
specialcalibrationrange(CalRange);andtheanalysisranges,ifmorethanone,
maybeprogrammedforseparateoridenticalgasapplications.
(2)Ifallrangesareforthesameapplication,thenyouwillnotneedtheCal
Range.Calibratinganyoneoftherangeswillautomaticallycalibratetheothers.
(3)If:a)eachrangeisprogrammedforadifferentgasapplication,b)your
sensorcalibrationhasdriftedlessthan10 %,andc)yourCalRangewascali-
bratedalongwithyourotherrangeswhenlastcalibrated,thenyoucanusethe
CalRangetocalibrateallapplicationsrangesatonce.
IfyourModel2010Banalyzerfitstheparagraph(3)description,above,
usetheCalRange.Ifyouranalyzerhasdriftedmorethan10 %,calibrateeach
rangeindividually.
CAUTION: Always allow 4-5 hours warm-up time before calibrat-
ing, if your analyzer has been disconnected from its
power source. This does not apply if the analyzer
was plugged in but was in STANDBY.
Teledyne Analytical Instruments
Part I 4-11
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
Theanalyzeriscalibratedusingreference,zero,andspangases.Gas
requirementsarecoveredindetailinchapter3,section3.4GasConnections.
Checkthatcalibrationgasesareconnectedtotheanalyzeraccordingtothe
instructionsinsection3.4,observingalltheprescribedprecautions.
Note: Shut off the gas pressure before connecting it to the analyzer,
and be sure to limit pressure to 40 psig or less when turning it
back on.
Readjustthegaspressureintotheanalyzeruntiltheflowratethroughthe
sensorsettlesbetween50to200cc/min(approximately0.1to0.4scfh).
Note: Always keep the zero calibration gases flow as close to the
flowrate of sample gas as possible
4.4.1 Zero Cal
TheZerobuttononthefrontpanelisusedtoenterthezerocalibration
function.Zerocalibrationcanbeperformedineithertheautomaticormanual
mode.
CAUTION: If you are zeroing the Cal Range by itself (multiple
application analyzers only), use manual mode
zeroing.
If you want to calibrate ALL of the ranges at once
(multiple application analyzers only), use auto mode
zeroing in the Cal Range.
Make sure the zero gas is flowing to the instrument. If you get a CELL
CANNOT BE BALANCED message while zeroing skip to section 4.4.1.3.
4.4.1.1
AutoModeZeroing
Observe the precautions in sections 4.4 and 4.4.1, above. PressZeroto
enterthezerofunctionmode.Thescreenallowsyoutoselectwhetherthezero
calibrationistobeperformedautomaticallyormanually. Usethe∆∇arrow
keys to toggle between AUTO and MAN zero settling. Stop when AUTO
appears,blinking,onthedisplay.
Selectzero
mode: AUTO
PressEntertobeginzeroing.
####.##% H2 N2
Slope=#.### CZero
4-12 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Thebeginningzerolevelisshownintheupperleftcornerofthedisplay.As
thezeroreadingsettles,thescreendisplaysandupdatesinformationonSlope=
inpercent/second(unlesstheSlopestartswithintheacceptablezerorangeand
doesnotneedtosettlefurther).Thesystemfirstdoesacoursezero,shownin
thelowerrightcornerofthescreenasCZero, for3 min, andthendoesafine
zero,anddisplaysFZero,for3 min.
Then,andwheneverSlopeislessthan0.01foratleast3min,insteadof
Slope you will see a countdown: 9 Left, 8 Left, and so fourth. These are
softwarestepsinthezeroingprocessthatthesystemmustcomplete,AFTER
settling, before it can go back toAnalyze. Software zero is indicated by S
Zero in the lower right corner.
####.## % H2N2
4Left=#.### SZero
Thezeroingprocesswillautomaticallyconcludewhentheoutputiswithin
theacceptablerangeforagoodzero.Thentheanalyzerautomaticallyreturnsto
the Analyze mode.
4.4.1.2
ManualModeZeroing
PressZerotoentertheZerofunction.Thescreenthatappearsallowsyou
toselectbetweenautomaticormanualzerocalibration.Usethe∆∇keysto
toggle between AUTO and MAN zero settling. Stop when MANUAL appears,
blinking,onthedisplay.
Selectzero
mode: MANUAL
PressEntertobeginthezerocalibration.Afterafewsecondsthefirstof
threezeroingscreensappears.Thenumberintheupperlefthandcorneristhe
first-stagezerooffset.Themicroprocessorsamplestheoutputatapredeter-
minedrate.
####.## % H2N2
Zeroadj:2048CZero
TheanalyzergoesthroughC–Zero,F–Zero,andS–Zero.DuringC–Zero
and F–Zero, use the ∆∇ keys to adjust displayed Zero adj: value as close as
possibletozero.Then,pressEnter.
S–Zerostarts.DuringS–Zero,theMicrocontrollertakescontrolasinAuto
ModeZeroing,above.Itcalculatesthedifferencesbetweensuccessivesam-
plingsanddisplaystherateofchangeasSlope= avalueinpartspermillionper
second(ppm/s).
Teledyne Analytical Instruments
Part I 4-13
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
####.## % H2 N2
Slope=#.### SZero
Generally, you have a good zero when Slope is less than 0.05 ppm/s for
about30seconds.
Oncezerosettlingcompletes,theinformationisstoredintheanalyzer’s
memory,andtheinstrumentautomaticallyreturnstotheAnalyzemode.
4.4.1.3 Cell Failure
Cellfailureinthe2010Bisusuallyassociatedwithinabilitytozerothe
instrumentwithareasonablevoltagedifferentialacrosstheWheatstonebridge.If
thisshouldeverhappen,the2010Bsystemalarmtrips,andtheVFDdisplaysa
failuremessage.
Cellcannotbebalanced
Checkyourzerogas
Beforereplacingthesensor:
a. Check your zero gas to make sure it is within specifications.
b. Check for leaks downstream from the sensor, where contamina-
tion may be leaking into the system.
c. Check flowmeter to ensure that the flow is no more than
200SCCM
d. Checktemperaturecontrollerboard.
e. Check gas temperature.
Ifnoneoftheaboveasindicated, thesensormayneedtobereplaced.
Checkwarranty,andcontactAnalyticalInstrumentsCustomerService.
4.4.2 Span Cal
TheSpanbuttononthefrontpanelisusedtospancalibratetheanalyzer.
Spancalibrationcanbeperformedineithertheautomaticormanualmode.
CAUTION: If you are spanning the Cal Range by itself (multiple
application analyzers only), use manual mode
zeroing.
If you want to calibrate ALL of the ranges at once
(multiple application analyzers only), use auto mode
spanning in the Cal Range.
Makesurethespangasisflowingtotheinstrument.
4-14 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
4.4.2.1
AutoModeSpanning
Observeallprecautionsinsections4.4and4.4.2, above. PressSpanto
enterthespanfunction.Thescreenthatappearsallowsyoutoselectwhetherthe
spancalibrationistobeperformedautomaticallyormanually. Usethe∆∇
arrow keys to toggle between AUTO and MAN span settling. Stop when
AUTO appears, blinking, onthedisplay.
Selectspan
mode: AUTO
PressEnterto move to the next screen.
SpanVal: 2Ø.ØØ%
<ENT>Tobeginspan
Usethe<>arrowkeystotogglebetweenthespanconcentrationvalue
andtheunitsfield(%/ppm).Usethe∆∇ arrowkeyschangethevalueand/orthe
units,asnecessary.Whenyouhavesettheconcentrationofthespangasyouare
using,pressEntertobegintheSpancalibration.
####.##% H2 N2
Slope=#.### Span
Thebeginningspanvalueisshownintheupperleftcornerofthedisplay.
Asthespanreadingsettles,thescreendisplaysandupdatesinformationon
Slope.Spanningautomaticallyendswhenthespanoutputcorresponds,within
tolerance,tothevalueofthespangasconcentration.Thentheinstrumentauto-
maticallyreturnstotheanalyzemode.
4.4.2.2
ManualModeSpanning
PressSpantostarttheSpanfunction.Thescreenthatappearsallows
youtoselectwhetherthespancalibrationistobeperformedautomaticallyor
manually.
Selectspan
mode: MANUAL
Use the ∆∇ keys to toggle between AUTO and MAN span settling. Stop
when MAN appears, blinking, on the display. Press Enterto move to the next
screen.
SpanVal: 2Ø.ØØ%
<ENT>Tobeginspan
Usethe<>arrowkeystotogglebetweenthespanconcentrationvalue
andtheunitsfield(%/ppm).Usethe∆∇ arrowkeyschangethevalueand/orthe
Teledyne Analytical Instruments
Part I 4-15
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
units,asnecessary.Whenyouhavesettheconcentrationofthespangasyouare
using,pressEntertobegintheSpancalibration.
PressEntertoenterthespanvalueintothesystemandbeginthespan
calibration.
Oncethespanhasbegun,themicroprocessorsamplestheoutputata
predeterminedrate.Itcalculatesthedifferencebetweensuccessivesamplings
anddisplaysthisdifferenceasSlopeonthescreen.Ittakesseveralsecondsfor
thefirstSlopevaluetodisplay.SlopeindicatesrateofchangeoftheSpan
reading.Itisasensitiveindicatorofstability.
####.##% H2N2
Slope=#.### Span
WhentheSpanvaluedisplayedonthescreenissufficientlystable,press
Enter.(Generally,whentheSpanreadingchangesby1 %orlessoftherange
beingcalibratedforaperiodoftenminutesitissufficientlystable.)OnceEnter
ispressed,theSpanreadingchangestothecorrectvalue.Theinstrumentthen
automaticallyenterstheAnalyzefunction.
4.5 The Alarms Function
TheModel2010Bisequippedwith6fullyadjustablesetpointsconcentra-
tionwithtwoalarmsandasystemfailurealarmrelay.Eachalarmrelayhasaset
ofform“C"contactsratedfor3amperesresistiveloadat250V ac.SeeFigure
inChapter3,Installationand/ortheInterconnectionDiagramincludedatthe
backofthismanualforrelayterminalconnections.
Thesystemfailurealarmhasafixedconfigurationdescribedinchapter3
Installation.
Theconcentrationalarmscanbeconfiguredfromthefrontpanelaseither
high or low alarms by the operator. The alarm modes can be set as latching or
non-latching, and either failsafe or non-failsafe, or, they can be defeated
altogether.Thesetpointsforthealarmsarealsoestablishedusingthisfunction.
Decidehowyouralarmsshouldbeconfigured.Thechoicewilldepend
uponyourprocess.Considerthefollowingfourpoints:
1. Which if any of the alarms are to be high alarms and which if any
are to be low alarms?
Setting an alarm as HIGH triggers the alarm when the
contaminant concentration rises above the setpoint. Setting an
alarm as LOW triggers the alarm when the contaminant
concentration falls below the setpoint.
Decide whether you want the alarms to be set as:
4-16 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
•
•
•
Both high (high and high-high) alarms, or
One high and one low alarm, or
Both low (low and low-low) alarms.
2. Are either or both of the alarms to be configured as failsafe?
In failsafe mode, the alarm relay de-energizes in an alarm
condition. For non-failsafe operation, the relay is energized in an
alarm condition. You can set either or both of the concentration
alarms to operate in failsafe or non-failsafe mode.
3. Are either of the alarms to be latching?
In latching mode, once the alarm or alarms trigger, they will
remain in the alarm mode even if process conditions revert back
to non-alarm conditions. This mode requires an alarm to be
recognized before it can be reset. In the non-latching mode, the
alarm status will terminate when process conditions revert to non-
alarmconditions.
4. Are either of the alarms to be defeated?
The defeat alarm mode is incorporated into the alarm circuit so
that maintenance can be performed under conditions which
would normally activate the alarms.
The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
Ifyouareusingpasswordprotection,youwillneedtoenteryourpassword
toaccessthealarmfunctions.Followtheinstructionsinsection4.3.3toenter
yourpassword.Onceyouhaveclearancetoproceed,entertheAlarmfunction.
Note: If all ranges are for the same application, setting one of them
will automaticaly set the others.
PresstheAlarmbuttononthefrontpaneltoentertheAlarmfunction.
Makesurethat01isblinking.
Selrngtosetalm:
> Ø1 Ø2 Ø3 <
SetuptheRange1alarmbymovingtheblinkingoverto01usingthe<>
arrowkeys.ThenpressEnter.Checkthegasapplicationandrangelimitsas
displayedonthescreen.
Gas use: H2 N2
Range: 0 10%
Pressenteragaintosetthealarmsetpoints.
Sel%/ppmalmtoset
AL1PPM AL2PPM
Teledyne Analytical Instruments
Part I 4-17
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
Use the ∆∇ keys to choose between % or ppm units. Then press Enter to
movetothenextscreen.
AL1: 1ØØØ ppm HI
Dft:N Fs:N Ltch:N
Fiveparameterscanbechangedonthisscreen:
•
•
•
•
•
Value of the alarm setpoint, AL1: ####
Out-of-range direction, HI or LO
Defeated? Dft:Y/N (Yes/No)
Failsafe? Fs:Y/N (Yes/No)
Latching? Ltch:Y/N (Yes/No).
•
•
To define the setpoint, use the < > arrow keys to move the
blinking over to AL1: ####. Then use the ∆∇ arrow keys to
change the number. Holding down the key speeds up the
incrementingordecrementing.
To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the ∆∇ arrow
keys to change the parameter.
•
•
Once the parameters for alarm 1 have been set, press Alarms
again, and repeat this procedure for alarm 2 (AL2).
To reset a latched alarm, go to Dft and then press either ∆ two
times or ∇ two times. (Toggle it to Y and then back to N.)
–OR –
Go to Ltch and then press either ∆ two times or ∇ two times.
(Toggle it to N and back to Y.)
4.6 The Range Select Function
TheRangefunctionallowsyoutomanuallyselecttheconcentration range
of analysis (MANUAL), or to select automatic range switching (AUTO).
In the MANUAL screen, you are further allowed to define the high and low
(concentration)limitsofeachRange,andselectasingle,fixedrangetorun.
CAUTION: If this is a linearized application, the new range must
be within the limits previously programmed using the
System function, if linearization is to apply through-
out the range. Furthermore, if the limits are too small
a part (approx 10 % or less) of the originally linear-
ized range, the linearization will be compromised.
4-18 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
IntheAUTOscreen,youarefurtherallowedtoselectwhichgasapplication
(PREVIOUSLYdefinedinSystemfunction)torun.
4.6.1 Manual (Select/Define Range) Screen
TheManualrange-switchingmodeallowsyoutoselectasingle,fixed
analysisrange.Itthenallowsyoutoredefinetheupperandlowerlimits,forthe
range.
PressRangekeytostarttheRangefunction.
Selectrange
mode: MANUAL
Note: If all three ranges are currently defined for different applica-
tion gases, then the above screen does not display (because
mode must be manual). Instead, the VFD goes directly to the
following screen.
If above screen displays, use the ∆∇ arrow keys to Select MANUAL, and
pressEnter.
Selectrangetorun
> Ø1 Ø2 Ø3 CAL<
Use the < > keys to select the range: 01, 02, 03, or CAL. Then press
Enter.
Gasuse: H2N2
Range: Ø 10 %
Usethe <> keystotogglebetweentheRange:low-endfieldandthe
Range:high-endfield.Usethe∆∇ keystochangethevaluesofthefields.
PressEscapetoreturntothepreviousscreentoselectordefineanother
range.
PressEntertoreturnthetotheAnalyzefunction.
4.6.2 Auto (Single Application) Screen
Autorangingwillautomaticalysettotheapplicationthathasatleasttwo
rangesetupwiththesamegases.
Intheautorangingmode,themicroprocessorautomaticallyrespondsto
concentrationchangesbyswitchingrangesforoptimumreadoutsensitivity.Ifthe
upperlimitoftheoperatingrangeisreached,theinstrumentautomaticallyshifts
tothenexthigherrange.Iftheconcentrationfallstobelow85%offullscaleof
Teledyne Analytical Instruments
Part I 4-19
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
thenextlowerrange,theinstrumentswitchestothelowerrange.Acorrespond-
ingshiftintheDCconcentrationoutput,andintherangeIDoutputs,willbe
noticed.
Theautorangingfeaturecanbeoverriddensothatanalogoutputstays ona
fixedrangeregardlessofthecontaminantconcentrationdetected.Iftheconcen-
trationexceedstheupperlimitoftherange,theDCoutputwillsaturateat1V dc
(20mAatthecurrentoutput).
However,thedigitalreadoutandtheRS-232outputoftheconcentration
areunaffectedbythefixedrange.Theycontinuetoreadbeyondthefull-scale
settinguntilamplifiersaturationisreached.Belowamplifiersaturation,the
overrangereadingsareaccurateUNLESStheapplicationuseslinearizationover
theselectedrange.
TheconcentrationrangescanberedefinedusingtheRangefunction
Manualscreen,andtheapplicationgasescanberedefinedusingtheSystem
function,iftheyarenotalreadydefinedasnecessary.
CAUTION: Redefining applications or ranges might require
relinearizationand/orrecalibration.
Tosetupautomaticranging:
PressRangekeytostarttheRangefunction.
Selectrange
mode : AUTO
Note: If all three ranges are currently defined for different applica-
tion gases, then the above screen does not display (because
mode must be manual).
If above screen displays, use the ∆∇ arrow keys to Select AUTO, and
press Enter.
PressEscapetoreturntothepreviousAnalyzeFunction.
4-20 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
4.6.3 Precautions
Part I: Control Unit
TheModel2010Ballowsagreatdealofflexibilityinchoosingrangesfor
automaticrangeswitching.However,therearesomepitfallsthataretobe
avoided.
Rangesthatworkwelltogetherare:
•
•
•
Ranges that have the same lower limits but upper limits that differ
by approximately an order of magnitude
Ranges whose upper limits coincide with the lower limits of the
next higher range
Ranges where there is a gap between the upper limit of the range
and the lower limit of the next higher range.
Rangeschemesthataretobeavoidedinclude:
•
•
Ranges that overlap
Ranges whose limits are entirely within the span of an adjoining
range.
Figure4-2illustratestheseschemesgraphically.
Figure 4-2: Examples of Autoranging Schemes
4.7 The Analyze Function
Normally,allofthefunctionsautomaticallyswitchbacktotheAnalyze
functionwhentheyhavecompletedtheirassignedoperations.Pressingthe
Teledyne Analytical Instruments
Part I 4-21
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
EscapebuttoninmanycasesalsoswitchestheanalyzerbacktotheAnalyze
function.Alternatively,youcanpresstheAnalyzebuttonatanytimetoreturn
toanalyzingyoursample.
TheAnalyzefunctionscreenshowstheimpurityconcentrationandthe
applicationgasesinthefirstline,andtherangeinthesecondline.Inthelower
rightcorner,theabbreviationAnlzindicatesthattheanalyzerisintheAnalyze
mode.Ifthereisan*beforetheAnlz,itindicatesthattherangeislinearized.
1.95 % H2 N2
nR1:Ø10 *Anlz
nindicatesnoninvertingrange
iindicatesinvertingrange
Iftheconcentrationdetectedisoverrange,thefirstlineofthedisplayblinks
continuously.
4.8 Programming
CAUTION: The programming functions of the Set Range and
Curve Algorithm screens are configured at the facto-
ry to the users application specification. These func-
tions should only be reprogrammed by trained,
qualified personnel.
Toprogram,youmust:
1. Enter the password, if you are using the analyzer’s password
protectioncapability.
2. Connect a computer or computer terminal capable of sending an
RS-232 signal to the analyzer RS-232 connector. (See chapter 3
Installation for details). Send the rp command to the analyzer.
OR
For software 1.1.4 or later, turn the instrument off and back on.
While on the introduction screen hold the Analyze key for at least
fifteen seconds. Press the Enter key twice to return to the
Analyze mode.
3. Press the System button to start the System function.
CONTRAST AUTOCAL
PWD LOGOUT MORE
ContrastFunctionisDISABLED
(Refer to Section 1.6)
Use the < > arrow keys to blink MORE, then press Enter.
4-22 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
MODEL APPLICATION
SELF_TEST ALGORITHM
NowyouwillbeabletoselecttheAPPLICATIONandALGORITHM
set-upfunctions.
4.8.1 The Set Range Screen
TheSetRangescreenallowsreprogrammingofthethreeanalysisranges
andthecalibrationrange(includingimpuritygas,backgroundgas,lowendof
range,highendofrange,and%orppmunits).Originalprogrammingisusually
doneatthefactoryaccordingtothecustomer’sapplication.Itmustbedone
throughtheRS-232portusingacomputerrunningaterminalemulationprogram.
Note: It is important to distinguish between this System programming
subfunction and the Range button function, which is an operator
control. The Set Range Screen of the System function allows the user
to DEFINE the upper and lower limits of a range AND the application
of the range. The Range button function only allows the user to
select or define the limits, or to select the application, but not
to define the application.
NormallytheModel2010Bisfactorysettodefaulttomanualrange
selection,unlessitisorderedasasingle-applicationmultiple-rangeunit(inwhich
caseitdefaultstoautoranging).Ineithercase,autorangingormanualrange
selectioncanbeprogrammedbytheuser.
Intheautorangingmode,themicroprocessorautomaticallyrespondsto
concentrationchangesbyswitchingrangesforoptimumreadoutsensitivity.Ifthe
upperlimitoftheoperatingrangeisreached,theinstrumentautomaticallyshifts
tothenexthigherrange.Iftheconcentrationfallstobelow85%offullscaleof
thenextlowerrange,theinstrumentswitchestothelowerrange.Acorrespond-
ingshiftintheDCconcentrationoutput,andintherangeIDoutputs,willbe
noticed.
Theautorangingfeaturecanbeoverriddensothatanalogoutputstays ona
fixedrangeregardlessofthecontaminantconcentrationdetected.Iftheconcen-
trationexceedstheupperlimitoftherange,theDCoutputwillsaturateat1V dc
(20mAatthecurrentoutput).
However,thedigitalreadoutandtheRS-232outputoftheconcentration
areunaffectedbythefixedrange.Theycontinuetoreadbeyondthefull-scale
settinguntilamplifiersaturationisreached.Belowamplifiersaturation,the
overrangereadingsareaccurateUNLESStheapplicationuseslinearizationover
theselectedrange.
Toprogramtheranges,youmustfirstperformthefourstepsindicatedat
thebeginningofsection4.8Programming.Youwillthenbeinthesecond
Systemmenuscreen.
Teledyne Analytical Instruments
Part I 4-23
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
MODEL APPLICATION
SELF_TEST ALGORITHM
Use the < > arrow keys again to move the blinking to APPLICATION
and press Enter.
Selrngtosetappl:
> Ø1 Ø2 Ø3 CAL <
Usethe∆∇arrowkeystoincrement/decrementtherangenumberto01,
02, 03, or CAL, and press Enter.
Imp: H2 Bck: N2
FR:Ø TO:1Ø %
Usethe<>arrowkeystomovetoImp:(impurity),Bck:(background),
FR:(from—lowerendofrange),TO:(to—upperendofrange),andPPMor%.
Usethe∆∇arrowkeystoincrementtherespectiveparametersasdesired.
Press Enter to accept the values and return to Analyze mode. (See note
below.)Repeatforeachrangeyouwanttoset.
Note: The ranges must be increasing from low to high, for example,
if Range 1 is set to 0–10 % and Range 2 is set to 0–100 %, then
Range 3 cannot be set to 0–50 % since that makes Range 3
lower than Range 2.
Ranges,alarms,andspansarealwayssetineitherpercentorppmunits,as
selectedbytheoperator,eventhoughallconcentration-dataoutputschange
fromppmtopercentwhentheconcentrationisabove9999 ppm.
Note: When performing analysis on a fixed range, if the concentra-
tion rises above the upper limit as established by the operator
for that particular range, the output saturates at 1 V dc (or 20
mA). However, the digital readout and the RS-232 output
continue to read regardless of the analog output range.
Toendthesession:
IfstartedwiththeRS-232,send:
st<enter>
st<enter>
totheanalyzerfromthecomputer.
Ifstartedthroughthefrontpanel,turntheinstrumentoffandbackon.
Press the Enter key twice to return to the Analyze mode.
4-24 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
4.8.2 The Curve Algorithm Screen
TheCurveAlgorithmisalinearizationmethod.Itprovidesfrom1to9
intermediatepointsbetweentheZEROandSPANvalues,whichcanbenormal-
izedduringcalibration,toensureastraight-lineinput/outputtransferfunction
throughtheanalyzer.
Eachrangeislinearizedindividually,asnecessary,sinceeachrangewill
usuallyhaveatotallydifferentlinearizationrequirement. Beforesettingthe
algorithmcurve,eachrangemustbeZeroedandSpanned.
Tolinearizetheranges,youmustfirstperformthefourstepsindicatedat
thebeginningofsection4.8Programming.Youwillthenbeinthesecond
Systemmenuscreen.
MODEL APPLICATION
SELF_TEST ALGORITHM
4.8.2.1 Checkingthelinearization
From the System Function screen, select ALGORITHM, and press
Enter.
Selrngsetalgo
> Ø1 Ø2 Ø3 <
Use the < > keys to select the range: 01, 02, or 03. Then press Enter.
Gasuse:H2 N2
Range: Ø-10 %
PressEnteragain.
Algorithmsetup:
VERIFY SETUP
Select and EnterVERIFY to check whether the linearization has been
accomplishedsatisfactorily.
Dpt INPUT OUTPUT
Ø Ø.ØØ Ø.ØØ
Theleftmostdigit(underDpt)isthenumberofthedatapointbeingmoni-
tored.Usethe∆∇keystoselectthesuccessivepoints.
TheINPUTvalueistheinputtothelinearizer.Itisthesimulatedoutputof
theanalyzer.Youdonotneedtoactuallyflowgas.
The OUTPUT value is the output of the linearizer. It should be the ACTU-
ALconcentrationofthespangasbeingsimulated.
Teledyne Analytical Instruments
Part I 4-25
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
IftheOUTPUTvalueshownisnotcorrect,thelinearizationmustbecorrect-
ed. Press ESCAPE to return to the previous screen. Select and Enter SET UP
toCalibrationModescreen.
Select algorithm
mode : AUTO
There are two ways to linearize: AUTO and MANUAL: The auto mode
requiresasmanycalibrationgasesastherewillbecorrectionpointsalongthe
curve.Theuserdecidesonthenumberofpoints,basedontheprecisionre-
quired.
Themanualmodeonlyrequiresenteringthevaluesforeachcorrection
pointintothemicroprocessorviathefrontpanelbuttons.Again,thenumberof
pointsrequiredisdeterminedbytheuser.
Note: Before performing section 4.8.2 or 4.8.2.3, you must check to
ensure that your calibration gases or points are between low
end and high end of the range setup. All correction points
must be between Zero and Span concentrations. Do not enter
Zero and Span points as part of the correction.
4.8.2.2 Manual Mode Linearization
Tolinearizemanually,youmusthavepreviousknowledgeofthenonlinear
thermal-conductivitycharacteristicsofyourgases.Youenterthevalueofthe
differentialbetweentheactualconcentrationandtheapparentconcentration
(analyzeroutput).AnalyticalInstrumentshastabulardataofthistypeforalarge
numberofgases,whichitmakesavailabletocustomersonrequest.SeeAppen-
dixfororderinginformation.Toenterdata:
From the System Functions Screen—
1. Use < > to select ALGORITHM , and Enter.
2. Select and Enter SETUP.
3. Enter MANUAL from the Calibration Mode Select screen.
Dpt INPUT OUTPUT
Ø Ø.ØØ Ø.ØØ
Thedataentryscreenresemblestheverifyscreen,butthegasvaluescan
bemodifiedandthedata-pointnumbercannot. Usethe<>keystotoggle
between the INPUT and OUTPUT fields. Use the ∆∇ keys to set the value for
thelowestconcentrationintothefirstpoint.ThenpressEnter.
4-26 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Aftereachpointisentered,thedata-pointnumberincrementstothenext
point.Movingfromthelowesttothehighestconcentration,usethe∆∇keysto
setthepropervaluesateachpoint.
Dpt INPUT
0 Ø.ØØ Ø.ØØ
OUTPUT
Repeattheaboveprocedureforeachofthedatapointsyouaresetting(up
toninepoints:0-8).Setthepointsinunitincrements.Donotskipnumbers.The
linearizerwillautomaticallyadjustforthenumberofpointsentered.
When you are done, Press ESCAPE. The message, Completed. Wait
for calculation, appears briefly, and then the main System screen returns.
Toendthesession:
IfstartedwiththeRS-232,send:
st<enter>
st<enter>
totheanalyzerfromthecomputer.
Ifstartedthroughthefrontpanel,turntheinstrumentoffandbackon.
Press the Enter key twice to return to the Analyze mode.
4.8.2.3 Auto Mode Linearization
TolinearizeintheAutoMode,youmusthaveonhandaseparatecalibra-
tiongasforeachofthedatapointsyouaregoinguseinyourlinearization.First,
theanalyzeriszeroedandspannedasusual.Then,eachspecialcalibrationgas,
foreachoftheintermediatecalibrationpoints,isflowed,inturn,throughthe
sensor.Aseachgasflows,thedifferentialvalueforthatintermediatecalibration
pointisenteredfromthefrontpaneloftheanalyzer.
Note: The span gas use to span the analyzer must be >90% of the
range being analyzed.
Beforestartinglinearization,performastandardcalibration.Seesection
4.4. Toenterdata:
From the System Functions screen—
1. Use < > to select ALGORITHM , and Enter.
2. Select and Enter SETUP.
3. Enter AUTO from the Calibration Mode Select screen.
TheAutoLinearizeModedataentryscreenappears.
1.95% H2 N2
Input(Ø):2.00
Teledyne Analytical Instruments
Part I 4-27
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
5. Use the ∆∇ keys to set the proper value of calibration gas, and
Enter. Repeat this step for each cal-point number as it appears in
the Input (x) parentheses.
6. Repeat step 5 for each of the special calibration gases, from the
lowest to the highest concentrations. Press Escape when done.
Toendthesession:
IfstartedwiththeRS-232,send:
st<enter>
st<enter>
totheanalyzerfromthecomputer.
Ifstartedthroughthefrontpanel,turntheinstrumentoffandbackon.
Press the Enter key twice to return to the Analyze mode.
4.9 Special Function Setup
4.9.1 Output Signal Reversal
Someapplicationsrequireareversaloftheoutputsignalsinorderforthe
4-20mAand0-1VDCoutputsignalstocorrespondwiththelowandhighend
oftheconcentrationrange. Forexample,ifanapplicationinvolvestheanalysis
of85%oxygeninabackgroundofargonbymeasuringthethermalconductivity
ofthebinarygas,theanalyzerwouldnormallybesetupsothatthe100%
oxygen(0%argon)concentrationwouldcorrespondtothezerolevel(4mA0V)
oftheoutputsignal. Then,85%oxygen(15%argon)wouldcorrespondto
20mA(1V)inthesignaloutput.
Itmaybeconvenientfortheusertohavetheoutputsreversedsothatthe
85-100%oxygenleveloutputsa4-20mA(0-1V)signalrespectively. Thiscan
beaccomplishedbyreversingthedatainputtothecustomsettings. Notall
applicationswillrequireareversingfunction,however,ifthisisdesirable,itmust
bespecifiedatthetimeofpurchaseoralternatively,bysubstitutingalinearizing
PCboardwiththereversalinformationcontainedtherein. Contactthefactory
forfurtherinformation.
4.9.2 Special - Inverting Output
NOTE: If the unit has a range or ranges that specified >0 unit setup
for inverting.
Thestepsare:
1.
Press RANGE key.
4-28 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
2.
UseLEFT/RIGHTkeytomovetotherangethatisspeci-
fiedasinvertingoutput.
3.
4.
PressandholdDOWNkeyforapproximately5to7seconds.
PressENTER key.
NOTE: If the inverting has been setup, “i” shall display on the left
bottom corner. Otherwise, the left bottom corner display ”n”.
Ifmorethatonerangeisspecifiedaninvertingoutput,repeatsteps1
to 4.
4.9.3 Special - Polarity Coding
NOTE: This setup will be identified only when performing GAS TEST
or calculation. Formula 1 will determine which range(s)
willrequired polarity coding.
Iftheanalyzerindicatesanegativeconcentrationwithproperspangasor
formula1indicatesanegativevalue,settheS1accordinglytothetablebelow:
Close S1-5 range 1
Close S1-6 range 2
Close S1-7 range 3
CloseS1-8calrange
PressI/Otorestartthesystem.
4.9.4 Special - Nonlinear Application Gain Preset
NOTE: This section apply during GAS TEST routine for the unit that
has more than one range install with nonlinear output applica-
tion.
Thestepsareasfollows:
1.
2.
Setunitrangetolowestrangereading.
Usingthecomputergeneratedsettingsforthecontroller,
adjustthecontrollersettingsforthemaximumspangas
output.
3.
PressSPANkey. SelectAUTOmodeandsetupthesetting
to span level. Press ENTER key to span.
4.
5.
Setrangeswitchtonextrange.
Press SPAN key.
Teledyne Analytical Instruments
Part I 4-29
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Operation
Model 2010B
6.
Press and hold the RIGHT key for approximate 5 to 7
seconds.
7.
SelectAUTOandsetthereadingtospangaslevel. Press
ENTERkey.
Repeat steps 1 to 7 if more than two ranges need to be setup.
4-30 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Maintenance
5.1 Routine Maintenance
Aside from normal cleaning and checking for leaks at the gas connec-
tions, routine maintenance is limited to replacing fuses, and recalibration. For
recalibration, see Section 4.4 Calibration.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.
5.2 System Self Diagnostic Test
1. Press the System button to enter the system mode.
2. Use the < > arrow keys to move to More, and press Enter.
3. Use the < > arrow keys to move to Self-Test, and press Enter.
The following failure codes apply:
Table 5-1: Self Test Failure Codes
Power
0
1
2
3
OK
5 V Failure
15 V Failure
Both Failed
Analog
0
1
2
3
OK
DAC A (0–1 V Concentration)
DAC B (0–1 V Range ID)
Both Failed
(Continued)
Part I 5-1
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
5 Maintenance
Model 2010B
Preamp
0
OK - > 0 means one or more gain have a
high offset, e.g. 204 means highest
gain has an offset. This error is
common on sealed air cells.
Cell
0
OK
1
Failed (open filament, short to ground, no
power.)
2
Unbalance(deteriorationoffilaments,blocked
tube)
5.3 VFD Display
NOTE:
When a vacuum Fluorescent Display is used, It will not
require contrast adjustment.
If you cannot read anything on the VFD, especially after first powering
up, check that VFD cable is not loose.
5.4 Fuse Replacement
The 2010B requires two 5 x 20 mm, 1 A, T type (Slow Blow) fuses.
The fuses are located inside the enclosure on the Electrical Connector Panel,
as shown in Figure 5-1. To replace a fuse:
1. Disconnect the Unit from its power source.
2. Place a small screwdriver in the notch in the fuse holder cap,
push in, and rotate 1/4 turn. The cap will pop out a few
millimeters. Pull out the fuse holder cap and fuse, as shown in
Figure 5-1.
5-2 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part I: Control Unit
Figure 5-1: Removing Fuse Cap and Fuse from Holder
3. Replace fuse by reversing process in step 1.
Remove Power to the instrument before changing fuses.
5.5 Major Internal Components
The Interconnection Panel and the Front Panel PCBs are accessed by
unlatching and swinging open the front door, as described earlier.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.
The 2010B Control Unit contains the following major components:
•
•
•
Power Supply
PreampandMotherboardwithMicrocontroller
Display Board and Displays
5 digit LED meter
2 line, 20 character, alphanumeric, VFD
Rear Panel Board.
•
Part I 5-3
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
5 Maintenance
Model 2010B
5.6 Cleaning
If instrument is unmounted at time of cleaning, disconnect the instru-
ment from the power source. Close and latch the front-panel access door.
Clean outside surfaces with a soft cloth dampened slightly with plain clean
water. Do not use any harsh solvents such as paint thinner or benzene.
For mounted instruments, DO NOT wipe the front panel while the
instrument is controlling your process. Clean the front panel as prescribed in
the above paragraph.
5.7 Phone Numbers
CustomerService: (626) 934-1673
Environmental Health and Safety: (626) 961-9221, Extension 1592
Fax: (626) 961-2538
EMERGENCY ONLY: (24-hour pager) 1-800-759-7243
PIN # 1858192
5-4 Part I
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
OPERATING INSTRUCTIONS
Model 2010B
Thermal Conductivity Analyzer
Part II: Analysis Unit
NEC Type
Part Number D-70089
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Model 2010B Thermal Conductivity Analyzer
Table of Contents
1 Introduction
1.1 Overview........................................................................ 1-1
1.2 Connections................................................................... 1-1
1.3 Electrical Connector Panel ............................................ 1-2
2 Installation
2.1 Unpacking the Analysis Unit.......................................... 2-1
2.2 Mounting the Analysis Unit ............................................ 2-1
2.3 Sample System Design ................................................. 2-2
2.4 Pressure and Flowrate Regulation ................................ 2-3
2.5 VENT Exhaust ............................................................... 2-3
2.6 SAMPLE Gas................................................................. 2-4
2.7 REFERENCE Gas......................................................... 2-4
2.8 ZERO Gas ..................................................................... 2-5
2.9 SPAN Gas...................................................................... 2-5
2.10 Electrical Connector Panel ............................................ 2-5
2.11 Signal Wiring Recommendations................................... 2-7
2.12 Testing the System......................................................... 2-7
3 Operation
3.1 System Self Diagnostic Test........................................... 3-1
3.2 Cell Failure Checks ....................................................... 3-2
4 Maintenance
4.1 Routine Maintenance..................................................... 4-1
4.2 Major Components......................................................... 4-1
4.3 Fuse Replacement......................................................... 4-2
4.4 System Self Diagnostic Test........................................... 4-3
4.5 Cell, Heater, and/or Thermistor Replacement ................ 4-4
4.5.1 Removing the Cell Compartment ........................ 4-4
4.5.2 Removing and Replacing the Cell Block............. 4-5
4.5.3 Removing the Heater and/or Thermocouple........ 4-6
4.5.4 Replacing the Heater and/or Thermister .............. 4-7
ii: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part II: Analysis Unit
Introduction
1.1 Overview
The Analytical Instruments Model 2010B Analysis Unit is a versatile
remotely controlled instrument for measuring a component gas in a back-
ground gas, or in a specific mixture of background gases.
Part 1 of this manual covers the Control Unit. Part II, this part, covers
the Model 2010B NEC type explosion proof Analysis Unit only.
1.2 Connections
The standard 2010 Analysis Unit is housed in a NEC type housing and
includes all gas connections. Figure 1-1 is a cutaway illustration of the
Analysis Unit showing the Gas Connections. The gas connectors are de-
scribed briefly here and in detail in the Installation chapter of this manual.
NOTE: The version of the 2010 Analysis Unit depicted has no options.
The gas connections found on the 2010B Analysis Unit depend on the
options selected. Refer to the outline diagram in the rear of this manual for
the actual configuration. These options may include gas panels with
flowmeters and auto-cal or manual valves as well as flame arrestors. If you
have purchased a unit with a sealed air option, there will not be any refer-
ence gas connections or flowmeter. If you have not purchased a unit with
gas panel, provision must be made to provide zero and span gas to the
sample in port.
•
SAMPLE IN This is the gas input from the sample stream.
This gas connection should have a flowmeter
with a range of 50-200 cc/min.
Teledyne Analytical Instruments
Part II: 1-1
Download from Www.Somanuals.com. All Manuals Search And Download.
1 Introduction
Model 2010B
Figure 1-1: Outline Diagram of 2010B Analysis Unit
•
REFERENCE IN This is the gas input from the flowing
reference gas source. This gas connector should
have a flowmeter with a range of 20-100 cc/
min.
•
•
SPAN/ZERO These gas inputs have the same requirements as
the sample in connection.
VENTS
Sample and reference gas vents must be re-
turned to areas of equal pressure.
The actual tubing size and connection type will depend on actual
options selected. See outline diagram in the rear of the manual for piping
requirements.
1.3 Electrical Connector Panel
Figure 1-2 shows the internal Electrical Connector Panel. Cables enter
the housing through access ports (visible in Figure 1-1), and connect to
terminals inside the housing. The connectors and controls are described
briefly here. They are described in detail in the Installation, Operation, and
Maintenancechapters, asappropriate.
1-2: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part II: Analysis Unit
Figure 1-2: Electrical Connector/Control Panel
•
Power In
Power input terminals for electric heater.
Requires 110 or 220 V ac, depending on
position of the Voltage Selector switch. Use
50/60 Hz.
CAUTION: Check the position of the Voltage Selector switch
BEFORE applying power to the Power Input termi-
nals.
•
VoltageSelector Power input selector switch for electric heater.
Adjusts input requirement for 110 or
220 V ac, depending on available source
voltage. Use 50/60 Hz.
•
Fuses
1.6 A, 250 V, T type, European size
5 × 20 mm fuses. Fuse 2 is on the neutral side
of the line. Fuse 1 is on the hot side of the
line.
•
•
Solenoid Valves Terminalsthatprovideallelectricalintercon-
nections from the Control Unit to the gas
controlvalves.
Sensor Signal
Terminals that provide connections from the
T/C sensor to the Control Unit.
Teledyne Analytical Instruments
Part II: 1-3
Download from Www.Somanuals.com. All Manuals Search And Download.
1 Introduction
Model 2010B
1-4: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part II: Analysis Unit
Installation
Installation of the Model 2010B Analyzer includes:
1. Unpacking, mounting, and interconnecting the Control Unit and
the Analysis Unit
2. Making gas connections to the system
3. Making electrical connections to the system
4. Testing the system.
2.1 Unpacking the Analysis Unit
The analyzer is shipped with all materials needed to install and prepare the
system for operation. Carefully unpack the Analysis Unit and inspect it for
damage. Immediately report any damage to the shipping agent.
2.2 Mounting the Analysis Unit
The Model 2010B Analysis Unit is for use in Class 1, Division 1,
Groups C and D, hazardous environments (group B available). The actual
class is limited by the flame arrestor option.
The standard model is designed for bulkhead mounting. Overall dimen-
sions of the Analysis Unit will vary depending on options. The maximum
footprint will be 19″ × 12″, and maximum height 9.4″. An Outline Drawing
at the back of this manual, gives the correct mounting dimensions for your
unit.
Note: The housing, including the cover, protrudes 8.6 to 9.4inches
from the base on which it is mounted. Enough clearance is
required in front of the cover to allow the cover to be re-
moved.
Figure 3-1 is a view with the cover removed showing the external Gas
Connector Panel and the internal Electrical Connector Panel.
Teledyne Analytical Instruments
Part II: 2-1
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Installation
Model 2010B
Figure 3-1: 2010 Analysis Unit with Auto-Cal & Gas Panel options
2-3 Sample System Design
Gas Connector and Selector Panels for specific applications are avail-
able at additional cost . These panels are designed to substitute a standard
frontpanel.
For those customers wishing to incorporate their own sample system,
electronic input/output ports are provided on the rear panel for the operation
of solenoid valves under the complete control of the Model 2010B electron-
ics. See section 3.3. The recommended system piping schematic is included
among the drawings at the rear of the manual
For best results, use the recommended piping system. Select a
flowmeter that can resolve 0.08 scfh (40-50 cc/min) for the reference path of
the analyzer, and select a flowmeter that can resolve 0.3 scfh (150 cc/min)
for the sample path of the analyzer.
2-2: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part II: Analysis Unit
NOTE:
NOTE:
The sample-line pressure regulator should be installed as
close to the sample point as possible to minimize sample-
line lag time.
An additional option is available for SEALED reference
application. This option would not have the reference Gas
Flow Meter, Piping and Fittings.
2-4 Pressure and Flowrate Regulation
Appropriate pressure reducing regulators must be installed at all gas
supply sources. To minimize flowrate adjustments the pressure regulators on
the supporting gas supply cylinders should be adjusted to provide the same
output pressure as the sample line regulator.
The gas pressure at the IN input should be reasonably well regulated.
Pressures between 5 and 50 psig are acceptable (10 psig is normal) as long
as the pressure, once established, will keep the flow constant during analy-
sis and within an acceptable range (between 0.1 and 0.4 scfh—See Note).
Note: Gases lighter than air have a flowrate higher than indicated on
the flowmeter, while gases heavier than air have a flowrate
lower than indicated. Values can range from one half to twice
the indicated flowrate.
For example: For hydrogen or helium, set the flowrate to 0.1
scfh (50 cc/min). For carbon dioxide or argon, set the flowrate
to 0.4 scfh (200 cc/min).
When installing pressure regulators on supply cylinders, crack the
cylinder valves so that gas is flowing during installation. This will eliminate
the most common cause of standardization-gas contamination: air trapped
during assembly diffusing back into the cylinder. This procedure is particu-
larly important in applications where impurity content of 1 to 2 % is the
range of interest.
2-5 VENT Exhaust
There are two separate VENT fittings—one for the sample gas and one
for the reference gas. Use 1/4 inch tubing for both sample and reference vents
to minimize back pressure from restricted flow.
Exhaust connections must be consistent with the hazard level of the
constituent gases. Check local, state, and federal laws, and ensure that the
exhaust stream vents to an appropriately controlled area if required. If not
vented to the same area, both VENT lines must vent to areas with equal
Teledyne Analytical Instruments
Part II: 2-3
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Installation
Model 2010B
ambient pressures, and pressures must vary no more than the normal baro-
metricchanges.
2-6 SAMPLE Gas
In the standard model, sample and calibration gases are introduced
through the SAMPLE fitting. The gases must be Tee'd into the Sample inlet
withappropriatevalves.
The gas pressure in should be well regulated. The sample line pressure
regulator should be installed as close to the sample line as possible to mini-
mize sample line lag time.
If greater flow is required for improved response time, install a bypass
in the sampling system upstream of the analyzer input.
2-7 REFERENCE Gas
A gas of fixed composition is needed as a reference to which the
sample gas will be compared. The reference gas is normally selected to
represent the main background gas of the analysis.
For most applications, a constant supply of reference gas flowing at the
same rate as the sample is required for best results. However, in many cases
the flow of reference gas can be slowed to about 0.08 scfh (40 cc/min) with
goodresults.
For some applications, an optional sealed air reference is installed. In
sealed-reference sensors the reference side of the detector cell is filled with
air and sealed. This eliminates the need to have reference gas constantly
passing through the cell.
NOTE:For instruments equipped with the optional sealed air refer-
ence, there is no REFERENCE inlet or reference VENT port.
It is highly recommended that the same cylinder of gas be used for both
the REFERENCE gas and the ZERO gas.
Pressure, flow, and safety considerations are the same as prescribed for
the SAMPLE gas, above.
2-4: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
2-8 ZERO Gas
Part II: Analysis Unit
For the ZERO gas, a supply of the background gas, usually containing
none of the impurity, is required to zero the analyzer during calibration. For
suppressed zero ranges the zero gas must contain the low-end concentration
of the impurity.
NOTE:Because most cylinder gases are between 99.95 and 99.98%
pure, it is highly recommended that the same cylinder of gas
be used for both REFERENCE and ZERO gas.
NOTE:It is essential to the accuracy of the analyzer that the purity of
the zero gas be known. Otherwise, when the zero control is
adjusted during zero standardization, the reading will indicate
the impurity content of the zero gas, rather than zero.
2-9 SPAN Gas
For the SPAN gas, a supply of the background gas containing 70-
100 % of the component of interest is required as a minimum.
Note: If your analyzer range is set for inverting output, your zero gas
will be at 100% of the range interest, and span will be 70 to
100% of the low end range.
If linearization is required, intermediate concentrations of the target gas
in the background gas may be necessary. From one to nine separate span
gases may be used, depending on the desired precision of the linearization.
See chapter 4, Operation.
2-10 Electrical Connector Panel
All electrical connections are made on the internal Electrical Connector
Panel, inside the explosion-proof enclosure, illustrated in Figure 3-3. The
signals are described in the following paragraphs. Wire size and length are
given in the Drawings section at the back of this manual. To access the
Panel, remove the explosion-proof cover as described in chapter 5, Mainte-
nance. NEVER OPEN THE COVER IN A HAZARDOUS ATMO-
SPHERE. THE AREA MUST BE DECLARED TEMPORARILY SAFE
BY THE PROPER AUTHORITY FIRST.
Teledyne Analytical Instruments
Part II: 2-5
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Installation
Model 2010B
Figure 3-4: Control Unit (CU) to Analysis Unit (AU) Connector
If you use your own gas control valves, use the interconnect diagram in
Figure 3-4 for the valves. The sensor and thermistor remain connected as in
Figure 3-5, above. (See drawing D-73170 for wire recommendations.)
Figure 3-5: Remote Probe Connector Pinouts
The voltage from the solenoid outputs is nominally 0 V for the OFF
and 15 V dc for the ON conditions. The maximum combined current that
can be pulled from these output lines is 100 mA. (If two lines are ON at the
same time, each must be limited to 50 mA, etc.)
If more current and/or a different voltage is required, use a relay, power
amplifier, or other matching circuitry to provide the actual driving current.
Note that each individual line has a series FET with a nominal ON resistance
of 5 ohms (9 ohms worst case). This can limit the obtainable voltage, de-
pending on the load impedance applied. See Figure 3-6.
2-6: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part II: Analysis Unit
Figure 3-6: FET Series Resistance
2-11 Signal Wiring Recommendations
The signal leads are the bias and sensor leads ref fig. 3.4, and reference
D-73170interconnectiondiagram. Themaximuminterconnectiondistances
between the 2010 control unit and the analysis unit is limited by the sensor
signal requirements. The sensor has two limiting factors. The first is series
resistance which is limited to a maximum of 2.5 ohms. Refer to the intercon-
nection diagram for wire gages and distances which will meet this require-
ment.
The second issue is one of signal quality. At greater distances prob-
lems associated with signal degradation increases. Signal degradation prob-
lems have two major issues. The first is associated with the local installation
environment. The number, proximity, and power associated with local
sources of EMI radiation can not be accounted for by TAI. These issues can
only be controlled by the end user. The other factor is the amplitude of the
signals generated by the thermal conductivity cell. This signal level is a
function of the gas composition of the gases involved. This signal level can
vary from micro-volts to mili-volts dependent on application. Consult TAI
for recommendations based on sample stream composition. Signal quality
and operating distances can be increased by the uses of better quality
shielded low loss cable. The cable configuration should be twisted pair with
double shielding for best performance. The signal cable should not be run
with AC power or digital signals.
2-12 Testing the System
After The Control Unit and the Analysis Unit are both installed and
interconnected, and the system gas and electrical connections are complete,
the system is ready to test. Before plugging either of the units into their
respectivepowersources:
Teledyne Analytical Instruments
Part II: 2-7
Download from Www.Somanuals.com. All Manuals Search And Download.
2 Installation
Model 2010B
•
•
•
Check the integrity and accuracy of the gas connections. Make
sure there are no leaks.
Check the integrity and accuracy of the electrical connections.
Make sure there are no exposed conductors
Check that sample pressure is between 3 and 40 psig, according
to the requirements of your process.
Power up the system, and test it as follows:
1. Repeat the Self-Diagnostic Test as described in Part I, chapter 4,
section 4.3.5.
2-8: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part II: Analysis Unit
Operation
3.1 System Self Diagnostic Test
The self diagnostics are run automatically by the analyzer whenever the
instrument is turned on, but the test can also be run by the operator at will.
During the test, internal signals are sent through the power supply, output
board and sensor circuit automatically. The return signal is analyzed, and at
the end of the test the status of each function is displayed on the screen,
either as OK or as a number between 1 and 3. (See Table 4-1 for number
code.)
Note: Remote Probe connector must be connected to the Analysis
Unit, or sensor circuit will not be properly checked.
Instructions for running self diagnostics are repeated here for your
convenience:
1. Press the System button to enter the system mode.
2. Use the < > arrow keys to move to More, and press Enter.
3. Use the < > arrow keys to move to Self-Test, and press Enter.
During preamp testing there is a countdown in the lower right corner of
the screen. When the testing is complete, the results are displayed.
Power:OK Analog:OK
Preamp:3
The following failure codes apply:
Table 4-1: Self Test Failure Codes
Power
0
1
OK
5 V Failure
Teledyne Analytical Instruments
Part II: 3-1
Download from Www.Somanuals.com. All Manuals Search And Download.
3 Operation
Model 2010B
2
3
15 V Failure
Both Failed
Analog
0
1
2
3
OK
DAC A (0–1 V Concentration)
DAC B (0–1 V Range ID)
Both Failed
Preamp
OK - > 0 means gains of the amplifier have a
high offset, e.g. 204 means the highest
0
gain has a high offset it is common on
sealedairreference.
The results screen alternates for a time with:
Press Any Key
ToContinue...
Then the analyzer returns to the initial System screen.
3.2 Cell Failure Checks
Cell failure is covered in detail in Part I: Control Units, section 4.4.1.3,
Cell Failure. Cell replacement is covered Part II: Analysis Units chapter 5,
Maintenance.
3-2: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part II: Analysis Unit
Maintenance
4.1 Routine Maintenance
Aside from normal cleaning and checking for leaks at the gas connec-
tions,routinemaintenanceislimitedtorecalibration.
Self-diagnostic testing of the system and fuse replacement in the Con-
trol Unit are covered in Part I, chapter 5 of this manual. For recalibration, see
Part I, section 4.4 Calibration.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.
4.2 Major Components
The internal components are accessed by rotating the explosion-proof
housing cover counterclockwise several turns until free. See Figure 4-1,
below.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
MANUAL.
The 2010B Analysis Unit contains the following major components:
•
•
•
•
•
TC Cell
Optional Auto-Cal Solenoid Assy.
HeaterAssembly
ElectricalConnectorPanel
Gas Connector Panel (external)
See the drawings in the Drawings section in back of this manual
fordetails.
Teledyne Analytical Instruments
Part II: 4-1
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Maintenance
Model 2010B
Figure 4-1: Major Components
4.3 Fuse Replacement
The 2010B Analysis Unit requires two 5 x 20 mm, 1.6 A, T type
(Slow Blow) fuses. The fuses are located inside the explosion proof housing
on the Electrical Connector Panel, as shown in Figure 4-2. To replace a fuse:
1. Disconnect the Unit from its power source.
2. Place a small screwdriver in the notch in the fuse holder cap,
push in, and rotate 1/4 turn. The cap will pop out a few
millimeters. Pull out the fuse holder cap and fuse, as shown in
Figure 4-2.
Figure 4-2: Removing Fuse Cap and Fuse from Holder
4-2: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part II: Analysis Unit
3. Replace fuse by reversing process in step 1.
4.4 System Self Diagnostic Test
1. Press the System button to enter the system mode.
2. Use the < > arrow keys to move to More, and press Enter.
3. Use the < > arrow keys to move to Self-Test, and press Enter.
4. Observe the error-code readings on the VFD Display screen, and
check Table 4-1, below, to interpret the codes.
Table 4-1: Self Test Failure Codes
Power
0
1
2
3
OK
5 V Failure
15 V Failure
Both Failed
Analog
0
1
2
3
OK
DAC A (0–1 V Concentration)
DAC B (0–1 V Range ID)
Both Failed
Preamp
OK - > 0 means gains of the amplifier have a
0
high offset, e.g. 204 means the highest
gain has a high offset it is common on
sealedairreference.
Teledyne Analytical Instruments
Part II: 4-3
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Maintenance
Model 2010B
4.5 Cell, Heater, and/or Thermistor
Replacement
The Thermal Conductivity Cell, with its Heater and Thermistor, is
mounted inside the insulated cell compartment, inside the analysis unit. To
remove these components, you must first remove screw on the cover, and
thecustomerinterface.
4.5.1 Removing the Cell Compartment
WARNING: IF THE MODEL 2010B ANALYZER HAS BEEN USED
WITH TOXIC GASES, FLUSH IT THOROUGHLY
BEFORE PERFORMING THIS PROCEDURE.
WARNING: DISCONNECT ALL POWER TO THE MODEL 2010B
BEFORE PERFORMING THIS PROCEDURE.
Remove the t/c cell as follows:
a. Verify that the power has been removed from the analysis unit
prior to removing cover.
b. Verify that the analysis unit is free from any toxic gas prior
disconnectinganyplumbing.
c. Remove the screw on lid and the six #6 screws securing the
interfacecoverpanel.
d. Remove the cover panel and the four 3/4" #6 stands-off securing
the 1st interface PCB.
e. Remove the 1st PCB and the four 7/16" #6 stand-off securing the
2nd interface PCB.
f. Unplug the heater from J3, and the sensor connectors from the
sensor connector J4, and J2, on the upper interface PCB.
Remove the 2nd PCB. Remove the two #10 screws from the
sensor retaining bracket located at the top of the sensor.
Disconnect the gas fitting and remove the sensor from the
analysisunit.
g. Reassemble in the reverse order, verify that there are no gas leaks
prior to returning the unit to service.
4-4: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part II: Analysis Unit
4.5.2 Removing and Replacing the Cell Block
a. Refer to Figure 4-3, which illustrates removal of the Cell Block
from the Cell Compartment. Exploded view is as seen from the
top of the Cell Block.
Figure 4-3: Removal of Cell from Cell Housing
Teledyne Analytical Instruments
Part II: 4-5
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Maintenance
Model 2010B
b. Remove the two screws holding the front mounting bracket—
they also hold the Cell Block Cover to the Cell Block—and then
pull off the cover.
c. Turn the uncovered Cell Block assembly over so that the bottom
faces you. The black rectangular block with four screws is the
Heater Block. Separate the Heater Block from the Cell Block by
removing the four screws. Leave the Heater Block electrical
connectionsconnected.
d. Remove the four screws from each of the black plates that hold
the Cell. The Cell is sandwiched between the plates. You should
now be able to slide the Cell free.
e. Leave the electrical connections connected at the Cell. Unlace the
cabling, and unplug the grey Cell cable at the Interface PCB
connector, J3. The Preamplifier PCB can be more easily
accessed by removing the analyzer's rear panel.
f. Replace the cell by reversing the above procedure, steps a
through e.
g. On analyzers with the sealed air reference option, the thermistor
is epoxied to the front of the cell block.
4.5.3 Removing the Heater and/or Thermocouple
a. Refer to Figure 4-4, which illustrates removal of the Thermistor
and/or Heater from the Cell Compartment. Exploded view is as
seen from the bottom of the Cell Block.
To Interface Board
Thermistor
Heater
Figure 4-4: Removing the Heater and/or Thermocouple
4-6: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Part II: Analysis Unit
b. Remove the two screws holding the front mounting bracket—
they also hold the Cell Block Cover to the Cell Block—and then
pull off the cover.
c. Turn the uncovered Cell Block assembly over so that the bottom
faces you. The black rectangular block with four screws is the
Heater Block.
d. The Heater is fastened to the Heater Block by a set screw as well
as the silicone sealing compound. The Thermistor is fastened
only by the silicone sealer. (NOTE: On units with sealed air
reference option, the thermistor is epoxied to the front on the
cell.)
(1) To remove the Heater, use a 1/16 ″ Allen wrench to loosen the
Thermistor set screw. Then, grasp BOTH Heater wires
firmly, and pull the Heater slowly out of the Heater Block,
breaking the silicone seal. Do not allow any foreign matter to
enter the empty duct.
(2) To remove the Thermistor, grasp BOTH Thermistor wires
firmly, and pull the Thermistor slowly out of the Heater
Block, breaking the silicone seal. Do not allow any foreign
matter to enter the empty duct.
e. Undo the cable lacing and separate the Heater/Thermistor wires.
Then, disconnect the wires from the interface PCB.
4.5.4 Replacing the Heater and/or Thermister
a. To replace the Heater and/or Thermister, coat the new element
with silicone sealing compound, and insert it into the duct.
CAUTION: The larger duct is for the Heater element, and the
smaller duct is for the Thermocouple.
b. Enough sealing compound should be on the element to spill over
and seal around the wire where it enters the duct. Smooth the
outer seal and remove any excess. (NOTE: On unit with the
sealed air reference option, the thermistor is expoxied to the
front of the t/c cell.)
c. Reassemble the Cell Compartment by reversing the procedure in
section 4.5.3. Then relace the cabling.
d. Reinstall the Assembled Cell Compartment by reversing the
procedure in section 4.5.1. Then reconnect the wires to J2 on the
interface PCB.
Teledyne Analytical Instruments
Part II: 4-7
Download from Www.Somanuals.com. All Manuals Search And Download.
4 Maintenance
Model 2010B
4-8: Part II
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
OPERATING INSTRUCTIONS
Models 2010B
Thermal Conductivity Analyzer
Appendix
Bulkhead Mount Control Unit, PN D70845
NEC Type Analysis Unit, PN D70089
NEC Type Analysis Unit, PN D70147
Teledyne Analytical Instrum
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix
Model 2010B
Contents
A-1 Model 2010B Specifications .......................................... A-3
A-2 Recommended 2-Year Spare Parts List ......................... A-5
A-3 Drawing List ................................................................... A-6
A-2
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Appendix
Appendix
A-1 Specifications
Ranges: Three ranges plus a cal range, field selectable
within limits (application dependent) and Auto
Ranging
Display: 2 line by 20 alphanumeric VFD accompanied
by 5 digit LED display
Accuracy: ±1% of full scale for most binary mixtures at
constanttemperature
±5% of full scale over operating temperature
range once temperature equilibrium has been
reached
Response Time: 90% in less than 50 seconds
SystemOperating
Temperature: 32°F to 122°F (0 - 50°C)
Sensor Type: StandardTCcell(4-filamentdetector)
SignalOutput: Two 0-1 VDC (concentration and range ID)
Two 4-20 mADC isolated (concentration and
range ID)
Alarm: Twofullyprogrammableconcentrationalarm
set points and corresponding Form C, 3 amp
contacts.
One system failure alarm contact to detect
power, calibration, zero / span and sensor
failure.
Teledyne Analytical Instruments
A-3
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix
Model 2010B
System Power
Requirements: Ex-Proof Analysis Unit 110 or 220 VAC, 50-
60Hz. (InternalSwitchselectable)
Control Unit: Universal Power 85-250 VAC 47-63Hz.
CellMaterial: Nickel plated brass block with nickel alloy
filaments and stainless steel end plates
O/PInterface: Full duplex RS-232, implement a subset of
Tracs Command
Mounting:
Standard: General purpose NEMA-4 Bulkhead mounted
Control Unit
Bulk Head mount Analysis Unit
Options: General purpose relay rack mounted to con-
tain either one or two in a 19” relay rack
mountableplate
* Other configurations, including a totally
explosion-proof,areavailable
Dimensions:
Control Unit: 15.75”H x 11.75”W x 8.5”D
Analysis Unit: 13.6”W x 18.1”H x 8.25”D
Control Unit: General Purpose Flush Panel Mounting
Analysis Unit: Explosion Proof Enclosure is UL and CSA
Listed for Class 1, Division 1, Groups B, C,
D Service. Nema 4/7 rated. (Actual Group
Rating is dependent on options selected)
Watted Parts: 316SS (except cell) standard version.
(Auto-Cal or Gas Panel option include viton
seals. The standard flowmeter includes viton
seals, SS valves, glass metering tube in an
acrylicblock).
A-4
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
Thermal Conductivity Analyzer
Appendix
A-2 Recommended 2-Year Spare Parts List
Qty
1
Part Number Description
D67472
Back Panel Board
1
C62371B
C65098A
C62365D
D70082
D67651
F768
Front Panel Board
1
PreamplifierBoard
1
Main Computer Board
1
Customer Interface PCB Assy (AU)
Std. T/C Cell Assy. (Brass + SS)
Fuse 1.6A (Analysis Unit)
2010 Front Door w/overlay
Fuse, 1 A, 250V, 5x20mm, T- Slo-Blo
Customer Interface PCB (CU)
1
2
1
D64849B
F1275
2
1
D65295B
_____________________
* Order one type only: US or European, as appropriate.
Note: Orders for replacement parts should include the part number
(if available) and the model and serial number of the instru-
ment for which the parts are intended.
Orders should be sent to:
TELEDYNE Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91749-1580
Phone (626) 934-1500
Fax (626) 961-2538
or your local representative.
Teledyne Analytical Instruments
A-5
Download from Www.Somanuals.com. All Manuals Search And Download.
Appendix
Model 2010B
A-3 Drawing List
C70381
D70117
C70146
C70335
D70081
D-73170
Outline Drawing (Standard & Auto-Cal option -C)
PipingDiagram
Outline Diagram (AU Auto-Cal Valve and Gas Panel -C-L)
Outline Diagram (gas panel option -L)
SchematicDiagram, InterfaceBoard
InterconnectionDiagram
A-6
Teledyne Analytical Instruments
Download from Www.Somanuals.com. All Manuals Search And Download.
|