Tektronix Switch VX4357 User Manual

User Manual  
VX4357  
Relay Switching Module  
070-9124-02  
This document supports firmware version 1.00  
and above.  
Warning  
The servicing instructions are for use by qualified  
personnel only. To avoid personal injury, do not  
perform any servicing unless you are qualified to  
do so. Refer to the Safety Summary prior to  
performing service.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contacting Tektronix  
Product  
Support  
For application-oriented questions about a Tektronix measure-  
ment product, call toll free in North America:  
1-800-TEK-WIDE (1-800-835-9433 ext. 2400)  
6:00 a.m. – 5:00 p.m. Pacific time  
Or contact us by e-mail:  
For product support outside of North America, contact your  
local Tektronix distributor or sales office.  
Service  
Support  
Contact your local Tektronix distributor or sales office. Or visit  
our web site for a listing of worldwide service locations.  
In North America:  
For other  
information 1-800-TEK-WIDE (1-800-835-9433)  
An operator will direct your call.  
To write us Tektronix, Inc.  
P.O. Box 1000  
Wilsonville, OR 97070-1000  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
EC Declaration of Conformity  
We  
Tektronix Holland N.V.  
Marktweg 73A  
8444 AB Heerenveen  
The Netherlands  
declare under sole responsibility that the  
VX4357  
meets the intent of Directive 89/336/EEC for Electromagnetic Compatibility and Low  
Voltage Directive 73/23/ECC for Product Safety. Compliance was demonstrated to the  
following specifications as listed in the Official Journal of the European Communities:  
EMC Directive 89/336/EEC:  
EN 55011  
Class A Radiated and Conducted Emissions  
EN 50081-1 Emissions:  
EN 60555-2  
AC Power Line Harmonic Emissions  
EN 50082-1 Immunity:  
IEC 801-2  
IEC 801-3  
IEC 801-4  
IEC 801-5  
Electrostatic Discharge Immunity  
RF Electromagnetic Field Immunity  
Electrical Fast Transient/Burst Immunity  
Power Line Surge Immunity  
Low Voltage Directive 73/23/EEC:  
EN 61010-1/A2 1995 Safety requirements for electrical equipment for  
measurement, control, and laboratory use  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix E – Performance Verification Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A – 25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
To avoid potential hazards, use this product only as specified.  
Installation Category Description  
CAT II LocalĆlevel mains (wall sockets).  
Equipment at this level includes appliances,  
portable tools, and similar products. Equipment  
is usually connected by a power cord.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CAT II, 48 V dc  
Download from Www.Somanuals.com. All Manuals Search And Download.  
V1.4  
V1.4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SA, 125V, fast blow  
Littelfuse  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
To avoid electric shock, tighten the module mounting screws after  
installing the module into the mainframe to ensure that the front panel is  
properly grounded.  
!
There are two labeled ejector handles on the card. To install the card  
correctly, make sure the ejector labeled “VX4357” is at the top.  
Note.  
Note.  
To maintain proper mainframe cooling, unused mainframe slots must be  
covered with the blank front panels supplied with the mainframe.  
Verify that the mainframe is able to provide adequate cooling and power  
with this module installed. Refer to the mainframe Operating Manual for  
instructions.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix E: Performance Verification Procedure  
The performance verification procedure for the VX4357 Relay Switching Module checks that the  
module is operating within specification.  
Conventions Used In This Procedure  
Please familiarize yourself with the following conventions which apply throughout this proce-  
dure:  
H
All control of the VX4357 Module will be accomplished through a VXI Slot 0 device. ASCII  
characters will form the commands sent to control the module and ASCII characters will be  
read from the module.  
H
Throughout this document writes to the VX4357 Module will be shown as: Send “Q1lf”  
Where Q1lf is the command (Query relay one) terminated with a linefeed character (lf).  
Commands must be sent exactly as shown.  
H
Reads from the module are shown as: Read “0 crlf”  
Where 0crlf is the ASCII string returned from the VX4357 Module, terminated with a carriage  
return and linefeed (crlf). The ASCII string returned from the module will be exactly as  
shown.  
H
H
Unless noted, all commands are sent to the VX4357 Module and all ASCII strings read are  
read from the VX4357 Module.  
Two characters are used throughout the communication sequences: “cr” is the carriage return  
character (character 13) and “lf” is the line feed character (character 10). These characters are  
the terminating characters for communication. They are shown throughout this document in  
the font shown here. Most reads from the VX4357 Module are terminated with both  
characters and are shown as “crlf”. Writes to the module require only a “lf”, but both  
characters may be used.  
H
An ASCII upper case alpha character “O” will be shown as “O”, and an ASCII numeric zero  
will be shown as “0”.  
A–15  
VX4357  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix E: Performance Verification Procedure  
Equipment Required  
Table 1–1 lists the equipment required for the performance and verification procedure.  
Table 1–1: Equipment Required  
Required tools and equipment  
Part number  
VXI Mainframe (such as the Tektronix VX1410)  
n/a  
n/a  
VXI Slot 0 with resource Manager (Tektronix VX4521) and appropriate cables and  
interface cards.  
Connector, DD-50P  
n/a  
n/a  
n/a  
n/a  
VX1785S connector  
Talker/Listener (Send/Read)  
4.5 digit Digital Multimeter with a 4-wire Ohms capability and capable of reading  
greater then 10 Gigohms.  
Connector Wiring  
Using the DD-50P connector wire the following pins together, leaving a pig tale, and label this  
N.O.:  
1, 3, 5, 7, 9, 11, 13, 15, 18, 20, 22, 24, 26, 28, 30, and 32  
Using the VX1785S connector wire the following pins together, leaving a pig tale, and label this  
N.C.:  
34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, and 49  
Default Conditions And LED Test  
1. Apply power to the VXI system and wait 10 seconds.  
Only the green PWR LED will be lit.  
2. Send “Slf” (Set all relays) to the VX4357.  
3. Verify that all LEDs for CH00 through CH31 are lit.  
4. Send “Rlf” (Reset all relays) to the VX4357.  
5. Verify that all LEDs for Ch00 through CH31 are out.  
DMM Connection And Set-Up  
Connect the DMM signal high and input positive (+) side to the pigtail on the DD-50P connector  
that is called out in the testing section.  
Connect the DMM signal low and input negative (–) side to the pigtail on the VX1780S connec-  
tor that is called out in the testing section. Set the DMM into 4 wire Ohms mode and autorange.  
A–16  
VX4357  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix E: Performance Verification Procedure  
Relay Resistance Test: Channel 00  
1. Place the DD-50P connector onto P4.  
2. Place the DMM + side to the pigtail labeled N.C., and place the DMM – side to DD-50P  
pin-2 (wiper Channel 00).  
3. The DMM will read less then 400 milliohms.  
4. Send “C0lf” (Close relay CH00) to the VX4357.  
5. The DMM will read greater then 10 Gigohms.  
6. Send “Q0lf” (Query relay CH00) to the VX4357.  
7. Read “1 crlf” (Relay CH00 is closed).  
8. Place the DMM + side to the pigtail labeled N.O .  
9. The DMM will read less then 400 milliohms.  
10. Send “O0lf” (Open relay CH00) to the VX4357.  
11. The DMM will read greater then 10 Gigohms.  
Repeat steps 1 through 11 for relays CH01 through CH15, substituting Cxlf and Oxlf for the ap-  
preciate relay, where x = relay to close or open. Use Table 1–2 for the channel to wiper pin-out.  
Table 1–2: Channel to wiper pin-out  
Channel  
01  
Wiper pin  
4
02  
6
03  
8
04  
10  
12  
14  
16  
19  
21  
23  
25  
27  
29  
31  
33  
05  
06  
07  
08  
09  
10  
11  
12  
13  
14  
15  
12. After repeating the above steps for Ch00 through CH15, remove the DD–50P connector from  
P4 and place it on P5.  
13. Add 16 to all relay references and repeat the above steps for relays CH16 through CH 31.  
A–17  
VX4357  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix E: Performance Verification Procedure  
Delay Command Test  
1. Send “Rlf” (Reset the VX4357) to the VX4357.  
2. Send “D1000C0C1” (Delay, Close CH00 and CH01 on the VX4357) to the VX4357.  
Monitor the time difference between CH00 LED to lit and CH01 to lit. This time will be  
approximately 1 second.  
3. Send “Tlf” (Delay Time of the VX4357) to the VX4357.  
4. Read “1000 crlf” (Time Delay programmed to 1000 milliseconds).  
A–18  
VX4357  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Toshiba DVD Player SD K740 User Manual
Toshiba Model Vehicle 23L1400UC User Manual
Toshiba Printer EO1 32004 User Manual
Traulsen Freezer TR35763 User Manual
Ultra Products Power Supply ULT31502 User Manual
Uniden Camera Accessories UDWC23 User Manual
Valor Auto Companion Inc Indoor Fireplace 1210DKA User Manual
Vector Camping Equipment VEC1198POB User Manual
ViewSonic Flat Panel Television VP171 User Manual
Walker Indoor Furnishings WQ42BC3BL User Manual