Omron Blood Pressure Monitor M1 Plus User Manual

Manual Inflation Blood Pressure Monitor  
Model M1 Plus  
• Instruction Manual  
• Mode d’emploi  
EN  
FR  
DE  
IT  
• Gebrauchsanweisung  
• Manuale di istruzioni  
• Manual de instrucciones  
• Gebruiksaanwijzing  
• ꢀYꢃTHTJWXHT ꢁT ꢂꢃWꢁQYFXF\NN  
• Instrukcja  
ES  
NL  
RU  
PL  
HU  
• Kezelési Útmutató  
IM-HEM-4011C-E-03-10/2011  
1649533-0C  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
Thank you for purchasing the OMRON M1 Plus Upper Arm  
Blood Pressure Monitor.  
The OMRON M1 Plus is a manual inflation blood pressure  
monitor, operating on the oscillometric principle. It measures  
your blood pressure and pulse rate simply and quickly.  
The unit also stores up to 21 measurements in memory and  
calculates an average reading based on the three most recent  
measurements taken within 10 minutes of the last reading. (If  
there are only two readings in memory for that period, the  
average will be based on two readings. If there is only one  
reading in memory for that period, that reading will be  
displayed as the average value.)  
Please read this instruction manual thoroughly before  
using the unit. For specific information about your own  
blood pressure, CONSULT YOUR DOCTOR.  
EN  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Important Safety Information  
Consult your doctor during pregnancy, arrhythmia and  
arteriosclerosis.  
You should never change the dosage of medication prescribed by  
your doctor.  
Use the M1 Plus only for yourself.  
Caution:  
• Do not leave the unit unattended with infants or persons who  
cannot express their consent.  
• Do not use the unit for any purpose other than measuring  
blood pressure.  
• Do not use a cellular phone near the unit. This may result in  
incorrect operation of the unit.  
• Do not operate unit in a moving vehicle (car, airplane).  
• Do not apply strong shocks and vibrations to or drop the main  
unit and arm cuff.  
• Only use four AAAmanganese or alkaline batteries with this  
unit. Do not use other types of batteries.  
• When using batteries, the battery liquid may leak and damage  
the main unit.  
• Please observe the following points.  
- Do not insert the batteries with their polarities incorrectly aligned.  
- Replace old batteries with new ones immediately.  
- When you are not going to use the unit for a long period of time  
(approximately three months or more), take out the batteries.  
- Do not use new and used batteries together.  
- Do not use batteries of a different type together.  
• Do not forcibly bend the arm cuff or bend the air tube  
excessively.  
• When removing the air tube, pull on the edge of the tube at  
the connection with the main unit not the middle of the tube.  
• Do not inflate the arm cuff when it is not wrapped around your  
arm.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. Overview  
Main unit  
A
F
E
D
B
C
C
A. Display  
F. Battery compartment  
B. SET button  
C. Air connector  
D. MEMORY button  
E. I/O button  
EN  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.Overview  
Arm Cuff  
G
H
G. Arm cuff  
(Medium cuff: arm circumference 22-32 cm)  
H. Air Tube  
If air starts to leak from the arm cuff, please replace with a new  
one. (Refer to Chapter 7 “Optional Parts”)  
Inflation Bulb  
J
I
K
I. Air Release Button  
J. Air Tube  
K. Air Inflation Bulb  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.Overview  
Display  
L
M
N
R
T
O
S
U
P
Q
V
W
L. Systolic blood pressure  
M. Diastolic blood pressure  
N. Memory symbol  
R. Heartbeat symbol  
1. (Flashes during measurement)  
2. (If flashing after measurement  
is complete, indicates blood  
pressure out of recommended  
range)  
(Displayed when viewing  
values stored in memory)  
O. Average value symbol  
(Displayed when viewing  
average value for the last three  
measurements)  
S. Pulse  
T. Deflation symbol  
U. Date/Time  
P. Buzzer symbol  
Q. Battery low symbol  
V. Irregular heartbeat symbol  
W. Reinflation symbol  
Package contents  
EN  
X
Y
X. Storage case  
Y. Four “AAA” alkaline (LR03)  
batteries  
• Instruction manual  
• Guarantee card  
• Blood pressure pass  
• Blood Pressure Monitor unit  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Preparation  
2.1 Installing/Replacing the Batteries  
1. Turn the main unit upside down.  
2. Slide the battery cover in the  
direction of the arrow while  
pressing the ribbed part of the  
cover.  
3. Install or replace four AAA”  
size batteries so that the  
+ (positive) and - (negative)  
polarities match the polarities  
indicated on the battery  
compartment.  
4. Put the battery cover back in  
place.  
Slide the battery cover as indicated, until it clicks into place.  
Note: The measurement values continue to be stored in  
memory (for maximum 30 seconds) even after the  
batteries are replaced.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.Preparation  
Battery Life & Replacement  
If the battery low symbol ( ) appears on the display, replace  
all four batteries at the same time.  
- When the battery low symbol ( ) starts to blink, you will still  
be able to use the unit for a short while. You should replace  
the batteries with new ones ahead of time.  
- When the symbol ( ) remains lit, the batteries are  
exhausted. You should replace the batteries with new ones  
at once.Turn the unit off before replacing the batteries.  
• Remove the batteries if the unit will not be used for three  
months or more.  
• If the batteries are removed for 30 seconds, the Date/Time  
setting will need to be reset. See “2.2 Setting the Date and  
Time” for details.  
• Dispose of batteries according to applicable local regulations.  
Four new AAAalkaline batteries will last for approximately  
1500 measurements, when used to take two measurements a  
day.  
EN  
Since the supplied batteries are for monitoring use only, they  
may have a shorter life and not last for 1500 measurements.  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.Preparation  
2.2 Setting the Date and Time  
Your blood pressure monitor automatically stores up to 21  
measurements values, including date and time of  
measurement, in its memory and calculates an average value  
based on the three most recent measurements taken within 10  
minutes of the last reading. (If there are only two readings in  
memory for that period, the average will be based on two  
readings. If there is only one reading in memory for that period,  
that reading will be displayed as the average value.)  
To make use of the memory and average values functions:  
• Set the unit to the correct date and time before taking a  
measurement for the first time.  
• If the batteries have been removed for more than 30 seconds,  
the date and time setting will need to be reset.  
1. When the I/O button is  
pressed to turn the unit on for  
the first time after inserting  
batteries, the year digits  
(2005) will flash on the  
display.  
2. Press the MEMORY button to  
advance the digits one at a  
time.  
Notes:  
• The range for the year  
setting is 2005 to 2030. If the  
year reaches 2030, it will  
return to 2005.  
• If you hold down the  
MEMORY button, the digits  
will advance rapidly.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.Preparation  
3. Press the SET button to confirm the setting when the desired  
number appears on the display.  
The year is set and the month digits flash on the display.  
4. Repeat steps 2 and 3 to set  
the month.  
The month is set and the day  
digits flash on the display.  
5. Repeat steps 2 and 3 to set  
the day.  
The day is set and the hour  
digits flash on the display.  
6. Repeat steps 2 and 3 to set  
the hour.  
The hour is set and the  
minutes digits flash on the  
display.  
EN  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.Preparation  
7. Repeat steps 2 and 3 to set  
the minutes.  
The minutes settings is set.  
The unit automatically  
switches to standby mode  
after the minute setting has  
been set.  
To adjust the date and time, press the SET button twice while  
the monitor is in standby mode.  
Notes:  
• After the buzzer setup is completed, the time/date setting will  
flash on the display.  
• After each setup is completed, please push the I/O button to  
turn off the monitor.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.Preparation  
2.3 Adjusting the Settings Modes  
You can modify the options for the various settings of your unit.  
This is done by pressing the SET button to select a setting,  
then pressing the MEMORY button to select the options for that  
setting. After selecting a setting, press the I/O button to confirm  
the setting and turn the power off.  
Buzzer Mode  
When Buzzer mode is on, the unit emits sounds, during  
measurement (to indicate that pulse waves are detected), and  
when measurement is complete.  
1. Press the SET button once  
while the unit is in standby  
mode.  
The buzzer symbol  
appears on the display.  
2. Press the MEMORY button to select  
onor off.  
The setting alternates between “on”  
and “off” with each press of the  
MEMORY button.  
EN  
The default setting is “on”.  
3. When you have selected the desired setting, press the I/O  
button to turn the unit off.  
Notes:  
• If you want to adjust the date or time, press the SET button  
after setting the buzzer setting, follow steps for setting date  
and time (Refer to 2.2) press the I/O button to turn the unit off.  
• If you forget to turn the unit off, it will automatically shut itself  
off after five minutes.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Using the Unit  
3.1 How to Sit Correctly When Taking a Measurement  
You can take a measurement on either your left or right arm.  
Note:  
• Measurements should be taken in a quiet place and you  
should be in a relaxed, seated position. Make sure that the  
room is not too hot or cold.  
• Do not take measurement after bathing, drinking alcohol, or  
exercising.  
• Do not move or talk during measurement.  
Correct  
Posture  
Remove tight fitting clothing from  
your upper arm, and any thick  
clothing such as a sweater.  
Do not place the cuff over thick  
clothes and do not roll up your sleeve  
if it is too tight.  
Sit upright  
with your  
back straight.  
Place your arm on a table so  
that the cuff will be at the same  
level as your heart.  
The gap between the  
chair and the top of the  
table should be between  
25cm to 30cm.  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.Using the Unit  
Notes:  
• Correct posture during measurement is necessary to get  
accurate results.  
You should also try to measure your blood pressure at the  
same time each day. (Within 1 hour after waking up is  
recommended.)  
Incorrect Posture  
Arched back (leaning  
forwards)  
Sitting cross-legged  
Sitting on a sofa or at a low  
table so that you lean forward  
These situations tend to lead to higher blood pressure values due to  
strain or the arm cuff being lower than the heart.  
If the arm cuff is at a lower position than your heart use cushions  
etc., to adjust the height of your arm.  
EN  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.Using the Unit  
3.2 Applying the Arm Cuff  
Be sure to wrap the arm cuff correctly so that you get accurate  
results.  
Measurements can be taken in light clothing. However, please  
remove thick clothes, such as sweaters, before taking a  
reading.  
1. Connect the air tube of the arm cuff and the inflation bulb to  
their respective air connectors.  
2. Put your left arm through the cuff  
loop.  
The air tube should run down the  
inside of your forearm and be in line  
with your middle finger.  
Note: If the cuff is not assembled,  
pass the end of the cuff  
furthest from the tubing  
through the metal D-ring to  
form a loop. The smooth cloth  
should be on the inside of the  
cuff loop.  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.Using the Unit  
3. Position the arm correctly.  
1) The air tube should run  
down the inside of your  
forearm and be in line with  
your middle finger.  
2) The bottom of the cuff  
should be approximately 1  
to 2 cm above your elbow.  
3) Apply the cuff to your  
upper arm so that the  
coloured marker (blue  
arrow under tube) is  
centered on the middle of  
your inner arm and points  
down the inside of the  
arm.  
Taking measurements on the right arm  
Apply the cuff so that the  
air tube is at the side of  
your elbow.  
• Be careful not to rest  
your arm on the air tube,  
or otherwise restrict the  
flow of air to the cuff.  
EN  
1 to 2 cm  
• The cuff should be 1 to  
2 cm above the elbow.  
4. When the cuff is positioned correctly, close the fabric fastener  
FIRMLY.  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.Using the Unit  
3.3 Taking a Reading  
1. Turning on the unit.  
Note: Press the air release button to release any air in the arm cuff.  
1) Press the I/O button to turn the unit on.  
2) All items in the display will be displayed briefly.  
The battery symbol will also be displayed for test purposes.  
This does not mean the batteries are empty.  
3) The current date is displayed and the deflation symbol will flash.  
4) Finally, the heartbeat symbol will appear and the current time  
and 0 is displayed.  
Note: If the deflation symbol does not disappear soon, press  
the air release button to release any air in the arm cuff.  
2. Pump the inflation bulb to inflate the arm cuff.  
1) Inflate the cuff until it is 30 to 40  
mmHg above your expected  
systolic blood pressure value.  
ex) If your expected blood  
pressure is around 140mmHg,  
inflate the arm cuff to between 170  
and 180 mmHg. Inflate the cuff  
rapidly so that the pressure is  
reached in about five seconds.  
2) When the desired pressure has been achieved, release the  
inflation bulb. Remain still and do not talk.  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.Using the Unit  
• The current time continues to be displayed while the cuff is  
being inflated.  
Note: Do not inflate the arm cuff more than necessary.  
3. Measurement starts.  
Measurement starts automatically  
after you stop inflating the arm cuff.  
Decreasing numbers appear on the  
display and the heartbeat symbol  
flashes.  
Remain still and do not talk during measurement.  
During measurement, the buzzer (if set to “on”) will beep in  
rhythm with your heartbeat.  
Note: To cancel a measurement, press the I/O button to turn  
off the unit and press the air release button to release  
the air in the arm cuff.  
Common Causes of Inaccurate Results  
Arm cuff is too loose.  
Movement or talking  
Leaning forwards.  
during measurement.  
EN  
Note: If the arm cuff is at a lower position than  
your heart use cushions etc., to adjust the  
height of your arm.  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.Using the Unit  
The deflation symbol flashes on the  
display. Blood pressure and pulse  
rate values are displayed.  
4. Measurement ends.  
Press the air release button to  
release the air in the arm cuff until  
the deflation symbol is no longer  
displayed.  
5. Check the measurement results.  
Systolic blood  
pressure  
The unit automatically stores blood  
pressure and pulse rate into its  
memory, including date and time of  
measurement. Refer to “3.4 Using  
the Memory Function”.  
Diastolicblood  
pressure  
Pulse  
The buzzer will sound (if switched  
on) when measurement is completed.  
Notes:  
• Self-diagnosis of measured results and treatment are  
dangerous. Please follow the instructions of your doctor.  
• Wait 2-3 minutes before taking another blood pressure  
measurement. Waiting between readings allows the arteries  
to return to the condition prior to taking the blood pressure  
measurement.  
6. Undo the fastener and remove the arm cuff.  
7. Press the I/O button to turn the unit off.  
Note: If you forget to turn the unit off, it will automatically shut  
itself off after five minutes.  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.Using the Unit  
Important:  
• If your systolic or diastolic pressure is outside  
the standard range, the heartbeat symbol will  
blink when the measurement result is  
displayed.  
Recent research suggests that the following  
values can be used as a guide to high blood  
pressure for measurements taken at home.  
Systolic Blood Pressure  
Above 135 mmHg  
Diastolic Blood Pressure  
Above 85 mmHg  
This criteria is for home blood pressure measurement.  
For professional office blood pressure measurement criteria,  
please refer to Chapter 9 “Some Useful Information about  
Blood Pressure”  
Your blood pressure monitor includes an  
irregular heartbeat feature. Irregular heartbeats  
can influence the results of the measurement.  
The irregular heartbeat algorithm automatically  
determines if the measurement is usable or  
needs to be repeated. If the measurement  
results are affected by irregular heartbeats but  
the result is valid, the result is shown together  
EN  
with the irregular heartbeat symbol (  
heartbeats cause the measurement to be invalid, no result is  
shown. If the irregular heartbeat symbol ( ) is shown after  
). If the irregular  
your have taken a measurement, repeat the measurement. If  
the irregular heartbeat symbol ( ) is shown frequently,  
please make your doctor aware of it.  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.Using the Unit  
What is Irregular Heartbeat?  
Normal Heartbeat  
Irregular Heartbeat  
An irregular heartbeat is a  
heartbeat rhythm that varies  
by more than 25% from the  
average heartbeat rhythm  
detected while the unit is  
measuring the systolic and  
diastolic blood pressure.  
Pulse  
Blood pressure  
Long  
Short  
Pulse  
If such an irregular rhythm is  
Blood pressure  
detected more than twice  
during measurement, the irregular heartbeat symbol (  
)
appears on the symbol when the measurement results are  
displayed.  
What is Arrhythmia?  
A heartbeat is stimulated by electrical signals that cause the  
heart to contract.  
Arrhythmia is a condition where the heartbeat rhythm is  
abnormal due to flaws in the bio-electrical system that drives  
the heartbeat. Typical symptoms are skipped heartbeats,  
premature contraction, an abnormally rapid (tachycardia) or  
slow (bradycardia) pulse. This can be caused by heart disease,  
aging, physical predisposition, stress, lack of sleep, fatigue etc.  
Arrhythmia can only be diagnosed by a doctor through a  
special examination.  
Whether the appearance of the irregular heartbeat symbol  
(
) in the results indicates arrhythmia or not can only be  
determined by an examination and diagnosis by your doctor.  
If the irregular heartbeat symbol (  
) is shown frequently,  
please make your doctor aware of it. Conducting self-diagnosis  
and treatment based on measurement results is dangerous. Be  
sure to follow the instructions of your doctor.  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.Using the Unit  
3.4 Using the Memory Function  
This unit has a memory capable of storing 21 sets of readings.  
Every time you complete the measurement, the unit  
automatically stores the blood pressure and pulse rate.  
If there are three readings within 10 minutes of the last reading,  
the average of the three readings is displayed. (If there are only  
two readings in memory for that period, the average will be based  
on two readings. If there is only one reading in memory for that  
period, that reading will be displayed as the average value.)  
Note: When 21 sets of readings are stored in memory, the  
oldest set will be deleted to store a new set.  
1. Press the O/I button to turn the  
power on.  
2. When the heartbeat symbol appears  
on the display, press the MEMORY  
button.  
EN  
If there are three readings stored in  
memory that were taken within 10  
minutes of the last reading, an  
average value for those reading is  
displayed. (If there are only two  
readings in memory for that period, the average will be based  
on two readings. If there is only one reading in memory for that  
period, that reading will be displayed as the average value.)  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.Using the Unit  
Note: If there are no measurements results stored in  
memory, the screen to the right is displayed.  
3. Press the MEMORY button again,  
while the average values screen is  
displayed.  
The result from the most recent  
measurement is displayed.  
The date and time of each  
measurement is displayed alternately  
with the measurement values.  
Important: If your systolic or diastolic pressure  
is outside the standard range, the  
heartbeat symbol will blink when  
the measurement result is  
displayed. Refer to 3.3, number 7.  
4. Press the MEMORY button  
repeatedly to cycle through the  
previous measurement results.  
Keep the button pressed down to  
cycle rapidly through the previous  
results.  
If the irregular heartbeat symbol was displayed at the time of a  
measurement, this is displayed when that result is displayed.  
5. Press the I/O button to turn the unit off.  
If you forget to turn the unit off, it will automatically shut itself off  
after five minutes.  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.Using the Unit  
To Delete All the Values Stored in Memory  
You cannot delete individual stored readings, all the readings in  
the unit will be deleted.  
1. To delete stored readings, first press  
the MEMORY button. Then while  
holding it down, press the I/O button  
simultaneously for about 2-3  
seconds. All readings will then be  
deleted.  
2
1
Important: Be careful not to press  
the I/O button first. If the  
I/O button is pressed  
first, the monitor is shut  
off.  
2. Press the I/O button to turn the unit off.  
If you forget to turn the unit off, it will automatically shut itself off  
after five minutes.  
EN  
Notes:  
• If the date and time are reset to a time before the most recent  
measurement, the average value will be based on any  
measurements taken after the date and time were reset.  
However, you can still view the readings stored in memory.  
• Please note that all measurements are stored in the memory. If  
different people measure with the same device, make sure that  
you are aware of this fact.  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Quick Reference Guide  
Use this as a quick reference guide only. If you are using this  
device for the first time, please read carefully Chapter 3 of this  
Instruction Manual.  
To help ensure a reliable reading, avoid eating, smoking, or  
exercising for at least 30 minutes before taking a measurement.  
Note: Remove any tight-fitting clothing from your upper arm.  
1. Sit on a chair with your  
feet flat on the floor and  
place your arm on a table  
so that the arm cuff will be  
at the same level as your  
heart.  
2. Apply the arm cuff to your upper arm. The coloured marker  
should be centred on the inside of your arm and point down the  
inside of the arm, so that the air tube runs down the inside of  
your forearm and is in line with your middle finger.  
3. Secure the cuff around your arm using the fabric fastener strip.  
4. Press the I/O button to turn the unit on, then  
pump the inflation bulb to inflate the cuff.  
After measurement is complete and the  
measurement results are displayed, press  
the air release button to release any air in  
the cuff.  
5. Press the I/O button to turn the unit off.  
Note: Always wait at least 2-3 minutes before taking another  
blood pressure measurement.  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Handling Errors and Problems  
5.1 Error Messages  
Error Display  
Cause  
Remedy  
Press air release button and  
restart the measurement with a  
higher inflation level. Carefully  
read steps under section 3.3.  
Cuff is under inflated.  
Repeat measurement. Remain  
still and do not talk during  
measurement.  
Movement during  
measurement  
Refer to section 3.3.  
Insert the air tube securely.  
Refer to section 3.2.  
Air tube disconnected.  
Apply the arm cuff correctly.  
Refer to section 3.2.  
Arm cuff not applied  
correctly.  
Remove any clothing interfering  
with the arm cuff.  
Refer to section 3.2.  
Clothing is interfering  
with the arm cuff.  
Replace cuff with new one.  
Refer to Chapter 7.  
Air is leaking from the  
arm cuff.  
EN  
Release your hand from the air  
inflation bulb before the pressure  
reaches 299 mmHg.  
The arm cuff was inflated  
above 299 mmHg.  
Refer to section 3.3.  
Replace all four “AAA” batteries  
with new ones.  
Refer to section 2.1.  
Battery power is low.  
Blinks or appears  
continuously  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.Handling Errors and Problems  
Error Display  
Cause  
Remedy  
Contact your OMRON retail outlet  
or distributor.  
Refer to Chapter 8.  
Device error.  
Press the inflation bulb to inflate  
the arm cuff until the reinflation  
symbol goes out. Or, deflate the  
arm cuff and repeat measurement  
after checking that the heartbeat  
symbol has been displayed.  
Pressure is too low.  
Refer to section 3.3.  
Note: The irregular heartbeat symbol may also be displayed with error messages.  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.Handling Errors and Problems  
5.2 Troubleshooting  
Problem  
Cause  
Remedy  
Apply the arm cuff  
correctly.  
Refer to section 3.2.  
Arm cuff not applied  
correctly.  
Remain still and do not talk  
during measurement.  
Refer to section 3.3.  
Movement or talking  
during measurement.  
The reading is  
extremely low (or high).  
Remove any clothing  
interfering with the arm  
cuff.  
Clothing is interfering  
with the arm cuff.  
Refer to section 3.1.  
Make sure that the air tube  
is connected securely.  
Refer to section 3.2.  
Is the air tube securely  
connected into the main  
unit?  
Arm cuff pressure does  
not rise.  
Replace the arm cuff with  
a new one.  
Refer to Chapter 7.  
Air is leaking from the  
arm cuff.  
Apply the cuff correctly so  
that it is firmly wrapped  
around the arm.  
Arm cuff deflates too  
soon.  
The arm cuff is loose.  
Refer to section 3.2.  
EN  
Inflate the cuff so that it is  
30 to 40 mmHg above  
your previous  
Has the arm cuff been  
inflated sufficiently?  
measurement result.  
Cannot measure or  
readings are too low or  
too high.  
Refer to section 3.3.  
Is the air release button  
being pressed during  
inflation?  
Be careful not to press the  
air release button during  
measurement.  
Replace the batteries with  
new ones.  
Refer to section 2.1.  
The unit loses power  
during measurement.  
The batteries are empty.  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.Handling Errors and Problems  
Problem  
Cause  
Remedy  
Replace the batteries with  
new ones.  
Refer to section 2.1.  
The batteries are empty.  
Nothing happens when  
you press the buttons.  
Insert the batteries with  
the correct (+/ -) polarity.  
Refer to section 2.1.  
The batteries have been  
inserted incorrectly.  
Press the I/O button and repeat measurement.  
If the problem continues, try replacing the batteries  
Other problems.  
with new ones.  
If this still does not solve the problem, contact your  
OMRON retail outlet or distributor.  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Maintenance and Storage  
Maintenance  
To protect your unit from damage, please observe the  
following:  
• Do not subject the main unit and cuff to extreme temperatures,  
humidity, moisture or direct sunlight.  
• Do not fold the cuff or tubing tightly.  
• Do not inflate the arm cuff over 299 mmHg.  
• Do not disassemble the unit.  
• Do not subject the unit to strong shocks or vibrations (for  
example, dropping the unit on the floor).  
• Do not use volatile liquids to clean the main unit. The unit should  
be cleaned with a soft, dry cloth.  
Use a soft, moistened cloth and soap to clean the arm cuff.  
• Do not wash the arm cuff or immerse it in water.  
• Do not use petrol, thinners or similar solvents to clean the arm  
cuff.  
EN  
• Do not carry out repairs of any kind yourself. If a defect  
occurs, consult your OMRON retail outlet or distributor as  
mentioned on the packaging.  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6.Maintenance and Storage  
Calibration and Service  
• The accuracy of this blood pressure monitor has been  
carefully tested and is designed for a long service life.  
• It is generally recommended to have the unit inspected every  
two years to ensure correct functioning and accuracy. Please  
consult your authorised OMRON dealer or the OMRON  
Customer Service at the address given on the packaging or  
attached literature.  
• Do not carry out any repairs yourself. If a defect occurs or you  
have doubts about the correct functioning of the device,  
consult your authorised OMRON dealer or the OMRON  
Customer Service.  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6.Maintenance and Storage  
Storage  
Keep the unit in its storage case when not in use.  
1. Unplug the air tube from the air connector.  
2. Gently fold the air tube into the arm  
cuff.  
Note: Do not bend the air tube  
excessively.  
3. Place the arm cuff, inflation bulb  
and main unit in the storage case.  
Do not store the unit in the following situations:  
• If the unit is wet.  
EN  
• Locations exposed to extreme temperatures, humidity, direct  
sunlight, dust or corrosive vapours.  
• Locations exposed to vibrations, shocks or where it will be at  
an angle.  
• Locations exposed to chemicals or corrosive vapours.  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Optional Parts  
Medium Arm Cuffs  
Large Arm Cuff  
Small Arm Cuff  
Arm circumference 17 - 22 cm  
Arm circumference 22 - 32 cm Arm circumference 32 - 42 cm  
CM-4997086-7  
CL-4997065-4  
CS-4997076-0  
CM1-7935058-8  
Small cuff and bulb combination  
Regular bulb  
Arm circumference 17 - 22 cm  
4997099-9  
4997965-1  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. Technical Data  
Product Description  
Model  
Display  
Manual Inflation Blood Pressure Monitor  
OMRON M1 Plus (HEM-4011C-E)  
LCD Digital Display  
Measurement Method  
Oscillometric method  
Pressure: 0 mmHg to 299 mmHg  
Pulse: 40 to 180/min.  
21 Measurements with date and time  
Measurement Range  
Memory  
Pressure: 3 mmHg  
Pulse: 5% of display reading  
Accuracy  
Inflation  
Manual by inflation bulb  
Deflation  
Pressure Detection  
Power Source  
Automatic pressure release valve  
Capacitive pressure sensor  
4 “AAA” batteries 1.5V  
Capacity of new alkaline batteries is approx. 1500  
Battery life  
measurements  
+10°C to +40°C  
Maximum: 30 to 85% RH  
Operating temperature/  
Humidity  
-20°C to +60°C  
Maximum: 10 to 95% RH  
Storage temperature  
Humidity  
Console Weight  
Cuff Weight  
Approximately 210g without batteries  
Approximately 130g  
Outer Dimensions  
Approximately 106 (w) mm 80 (h) mm 136 (l) mm  
Approximately 146 mm 446 mm  
Cuff Dimensions  
Package Content  
(Medium cuff: arm circumference 22 to 32 cm)  
Medium cuff, instruction manual, storage case, battery set,  
guarantee card, blood pressure pass, regular bulb.  
EN  
Note: Subject to technical modification without prior notice  
• This OMRON product is produced under the strict quality system of OMRON Healthcare  
Co. Ltd., Japan. The Core component for OMRON blood pressure monitors, which is the  
Pressure Sensor, is produced in Japan for assembly.  
• Disposal of this product and used batteries should be carried out in accordance  
with the national regulations for the disposal of electronic products.  
= Type B  
This device fulfils the previsions of EC directive 93/42/EEC (Medical Device  
Directive). This blood pressure monitor is designed according to the European Standard  
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part 3:  
Supplementary requirements for electromechanical blood pressure measuring systems.  
Caution: Please read the instruction manual carefully  
before using the device.  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8.Technical Data  
Important information regarding Electro Magnetic Compatibility (EMC)  
With the increased number of electronic devices such as PC’s and mobile (cellular)  
telephones, medical devices in use may be susceptible to electromagnetic  
interference from other devices. Electromagnetic interference may result in  
incorrect operation of the medical device and create a potentially unsafe situation.  
Medical devices should also not interfere with other devices.  
In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with  
the aim to prevent unsafe product situations, the EN60601-1-2 standard has been  
implemented. This standard defines the levels of immunity to electromagnetic  
interferences as well as maximum levels of electromagnetic emissions for medical  
devices.  
This medical device manufactured by OMRON Healthcare conforms to this  
EN60601-1-2:2001 standard for both immunity and emissions.  
Nevertheless, special precautions need to be observed:  
• Do not use mobile (cellular) telephones and other devices, which generate strong  
electrical or electromagnetic fields, near the medical device. This may result in  
incorrect operation of the unit and create a potentially unsafe situation. Recom-  
mendation is to keep a minimum distance of 7 m. Verify correct operation of the  
device in case the distance is shorter.  
Further documentation in accordance with EN60601-1-2:2001 is available at  
OMRON Healthcare Europe at the address mentioned in this instruction manual.  
Correct Disposal of This Product  
(Waste Electrical & Electronic Equipment)  
This marking shown on the product or its literature, indicates that it should not be  
disposed of, with other household wastes at the end of its working life. To prevent  
possible harm to the environment or human health from uncontrolled waste  
disposal, please separate this from other types of wastes and recycle it responsibly  
to promote the sustainable reuse of material resources.  
Household users should contact either the retailer where they purchased this  
product, or their local government office, for details of where and how they can take  
this item for environmentally safe recycling.  
Business users should contact their supplier and check the terms and conditions of  
the purchase contract. This product should not be mixed with other commercial  
wastes for disposal.  
This product does not contain any hazardous substances.  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9. Some Useful Information about Blood Pressure  
What is Blood Pressure?  
Blood pressure is a measure of the force of blood flowing  
against the walls of the arteries. Arterial blood pressure is  
constantly changing during the course of the heart’s cycle.  
The highest pressure in the cycle is called the Systolic Blood  
Pressure; the lowest is the Diastolic Blood Pressure.  
Both pressure readings, the Systolic and Diastolic, are  
necessary to enable a doctor to evaluate the status of a  
patient’s blood pressure.  
Why is it a Good Thing to measure Blood Pressure at  
Home?  
Having your blood pressure measured by a doctor can cause  
anxiety which is itself a cause of high blood pressure. As a  
variety of conditions affect blood pressure, a single  
measurement may not be sufficient for an accurate diagnosis.  
Many factors such as physical activity, anxiety, or the time of  
day, can influence your blood pressure. Thus it is best to try  
and measure your blood pressure at the same time each day,  
to get an accurate indication of any changes in blood pressure.  
Blood pressure is typically low in the morning and increases  
from afternoon to evening. It is lower in the summer and higher  
in the winter.  
EN  
Blood pressure is measured in millimetres of mercury (mmHg)  
and measurements are written with the systolic pressure  
before the diastolic e.g. A blood pressure written as 135/85, is  
referred to as 135 over 85 mmHg.  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9.Some Useful Information about Blood Pressure  
Upper curve: systolic blood pressure  
Lower curve: diastolic blood pressure  
200  
150  
100  
50  
6
12  
18  
24  
Time of day  
Example: fluctuation within a day (male, 35 years old)  
Classification of Blood Pressure by the World Heath  
Organization  
The World Health Organization (WHO) and the International  
Society of Hypertension (ISH) developed the Blood Pressure  
Classification shown in this figure.  
Systolic blood pressure  
(mmHg)  
Severe hypertension  
180  
Moderate hypertension  
160  
Mild hypertension  
140  
Normal systolic value  
130  
Normal blood  
pressure  
120  
Optimal blood  
pressure  
(target value)  
Diastolic  
blood  
pressure  
(mmHg)  
80  
85  
90  
100 110  
This classification is based on the blood pressure values  
measured on people in a sitting position in outpatient  
departments of hospitals.  
* There is no universally accepted definition of hypotension.  
However, those having the systolic pressure below  
100 mmHg are assumed as hypotensive.  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9.Some Useful Information about Blood Pressure  
Manufacturer  
OMRON HEALTHCARE Co., Ltd.  
53, Kunotsubo, Terado-cho, Muko, Kyoto,  
617-0002 JAPAN  
EU-representative  
OMRON HEALTHCARE EUROPE B.V.  
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS  
OMRON (DALIAN) CO., LTD.  
Production facility  
Subsidiary  
Dalian, CHINA  
OMRON HEALTHCARE UK LTD.  
Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, U.K.  
OMRON MEDIZINTECHNIK  
HANDELSGESELLSCHAFT mbH  
John-Deere-Str. 81a, 68163 Mannheim, GERMANY  
OMRON SANTÉ FRANCE SAS  
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE  
Made in China  
EN  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  

MRV Communications Server LX 4000 Series User Manual
NEC Cell Phone NDA 24300 User Manual
NHT Speaker S 00 Powered User Manual
Nokia Cell Phone X500 User Manual
Nostalgia Electrics Popcorn Poppers HKP200 User Manual
Omega Vehicle Security Electric Heater DRF POT User Manual
Onkyo Stereo Receiver TX 8222 User Manual
Panasonic Air Conditioner CW XC125HU User Manual
Panasonic Network Card 2SC5846G User Manual
Patton electronic Network Router 1000RC User Manual