Omega Engineering Sander CL531 User Manual

®
Users Guide  
Shop online at  
omega.com  
For latest product manuals:  
omegamanual.info  
CL531  
mA Loop Calibrator  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Keypad Operations  
1
EZ-Check™ Switch  
2
SOURCE/OFF/READ Switch  
Slide the SOURCE/OFF/READ Switch to SOURCE to  
output a mA signal and to do 2 - wire transmitter  
simulation. Use the READ position to read mA signal and  
power & measure 2 – wire transmitter.  
SOURCE mode: Slide the switch to select from Hi and low range pre-  
set values and the mid ranger (Dial) is selectable. Dial the mid range  
value and it will store the value with-in 5 seconds automatically.  
READ mode: Slide the switch to recall minimum and maximum  
readings. Press the EZ-Dial™ Knob to clear the stored values.  
EZ-Dial™ Knob  
Turn the knob to change display in 0.01mA increments.  
Push and turn for faster dialing. Push without turning to  
clear EZ-Check™ HI/LO points in READ mode.  
Press twice to select options:  
In Source mode select –  
% or mA  
2-Wire Transmitter Simulate  
% or mA  
CL531  
low power (15V) or High power (24V)  
mA LOOP CALIBRATOR  
In Read mode select -  
% or mA  
Power and Measure 2-Wire Transmitter  
% or mA  
low power (15V) or High power (24V)  
HART® Protocol  
An internal jumper enables the Power & Measure 2 – wire  
transmitter mode to be compatible with HART®  
communicators and transmitters.  
EZ-Dial™ Knob  
Adjust the output up and down with the EZ-Dial™ knob. The increment is 0.01 mA (or 0.1 % if display units are % of 4-20 mA.)  
Press while turning to adjust 10X faster – 0.10 mA (or 1.00 %.)  
Quick Reference Bar Graph  
The Quick Reference Bar Graph indicates the input and output level to the CL531 in % of 4-20 mA with 1.0% resolution. If the  
input or output signal is outside the normal operating range of the CL531 the Quick Reference Bar Graph in source mode will  
flash, in Read mode display over range when above 24.5mA.  
Error Conditions  
Bar Graph will flash when any error conditions exist.  
HART® Protocol  
Remove the back of the case and remove the jumper that is located in position J6 on the PC board. By doing so it places a 250Ω  
resister in series with the output of the CL531. This internal resister eliminates the need to add an external load resister when  
communicating with a HART® transmitter. This reduces the typical drive capability to 950 Ω.  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
EZ-Check  
Switch  
The EZ-Check™ switch has three positions -- high, dial, and low. Its position is shown at the left edge of the display with “HI” and  
“LO” indicators. Neither indicator indicates the middle position. Use of the EZ-Check™ switch depends on mode.  
Source Modes:  
Slide the EZ-Check™ switch to the HI and LO positions to recall the preset settings (Hi=20.00mA & Lo=4.00mA).  
Hint: For faster calibrations, the position of the switch can be felt. This feature allows continuous monitoring of the device  
being calibrated without looking back at the CL531 display. This is also useful in poor lighting or under difficult operating  
conditions.  
Read Modes:  
In read mode, the CL531 calibrator records the maximum and minimum readings observed in each mode. Slide the EZ-  
Check™ switch to the Dial position to read the loop. Then Slide the EZ-Check™ switch to the HI and LO positions to display  
to clear the readings. The display will flash “CLEARED” to confirm.  
knob  
EZ-Dial  
the max and min readings. Press  
Source Mode  
Receiver  
Transmitter Input  
+
IN  
-
REF +OUT-  
(Sensor)  
(Process Signal)  
(Simulated Input)  
Typical  
2-Wire  
Transmitter  
CL531  
mA LOOP CALIBRATOR  
Power Supply  
0 to 60 VDC  
Source mode uses internal power to supply current from 0.00-24.00 mA into as much as 1200Ω until the end of battery life. The  
calibrator Graph will flash if connected improperly. The three-position EZ-Check switch provides instant preset 4mA at zero, 12mA at  
mid range and 20mA at full scale outputs. The output is adjusted in 0.01 or 0.10 mA increments (0.1 or 1.0) % display units with  
the EZ-Dial knob.  
2-Wire Transmitter Simulation Mode  
Receiver  
Transmitter Input  
+
IN  
-
REF +OUT-  
(Sensor)  
(Process Signal)  
(Simulated Input)  
Typical  
2-Wire  
Transmitter  
CL531  
CL531 Pocket-Mate  
mA LOOP CALIBRATOR  
Power Supply  
to 60 VDC  
0
The CL531 can simulate a 2-wire transmitter in the 4-20 mA or % process loop. In source mode press the EZ-Dial™ Knob twice  
to get into the feature options. Then press the EZ-Dial™ Knob to select mA 2 – wire. The EZ-Check switch and EZ-Dial knob allow  
rapid and fine control of loop current.  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
Power and Measure Transmitter Mode  
Receiver  
Transmitter Input  
+
IN  
-
REF +OUT-  
(Sensor)  
(Process Signal)  
(Simulated Input)  
Typical  
2-Wire  
Transmitter  
CL531  
mA LOOP CALIBRATOR  
Power Supply  
0 to 60 VDC  
The CL531 supplies 15Volts or 24 Volts to the transmitter and displays the output in mA or % on the CL531 display. In read mode  
press the EZ-Dial™ Knob twice to get into the feature options. Then press the EZ-Dial™ Knob to select mA PWR - M. Then turn  
EZ-Dial™ Knob to select power range (15V or 24V). The EZ-Check switch and EZ-Dial knob allow rapid and fine control of loop  
current.  
Read Mode  
Receiver  
Transmitter Input  
+
IN  
-
REF +OUT-  
(Sensor)  
(Process Signal)  
(Simulated Input)  
Typical  
2-Wire  
Transmitter  
CL531 Poc  
CL531  
ket-Mate  
mA LOOP CALIBRATOR  
Power Supply  
0 to 60 VDC  
The CL531 can read loop currents from 0-24 mA. The CL531 limits current in read mode to 25mA to protect the devices in  
the loop from over voltage or over current conditions.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
General Specifications:  
(Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration)  
Operating Temperature Range  
Storage Temperature Range  
Relative Humidity Range  
-20 to 60 °C (-5 to 140 °F)  
-30 to 60 °C (-22 to 140 °F)  
10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing  
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing  
2 AA Alkaline  
Battery  
Miscellaneous  
Low battery indication with nominal 1 hour of operation left  
Over-voltage protection to 120 Vrms (rated for 30 seconds) or 240 Vrms (rated for 15  
seconds)  
Bar graph display with 1% resolution of 4-20 mA signal scale  
High contrast graphic liquid crystal display with 0.45” (11.4 mm) high digits  
Common Specifications for all current modes  
Ranges  
0.00 to 24.00 mA, 25.0 to 125.0% of 4-20 mA  
Accuracy  
(0.05 % of Reading + 0.01 mA)  
50 ppm/°C of Range  
Temperature effect  
Resolution(s)  
0.01 mA and 0.1 %  
Source/Power and Measure 2-Wire Transmitter Specifications:  
Loop compliance voltage  
Loop drive capability  
≥ 15 Volts or ≥ 24 Volts  
1200 Ω at 20 mA for entire battery life @ 24 Volts  
600 Ω at 20 mA for entire battery life @ 15 Volts  
Read mA Specifications:  
Voltage burden  
≤ 1V at 20 mA  
Overload/Current limit protection  
Battery life  
nominal ≤24 mA  
Typical  
40 Hours  
2-Wire Transmitter Simulation Specifications:  
Voltage burden  
≤ 2V at 20 mA  
nominal ≤ 24 mA  
Overload/Current limit protection  
Loop voltage limits  
Miscellaneous  
2-42 VDC  
Open loop or out of compliance conditions are indicated by appropriate error display  
Battery life ≥ 40 hour typical  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
CL531 Field Calibration Procedure  
1. Precision Current Meter with accuracy of 0.025% at 20mA.  
2. Precision Current Source with accuracy of 0.025% at 20mA  
Place in fresh batteries and allow one hour for the CL531 to stabilize to the ambient  
temperature. Remove the battery cover and the Phillips head screws. While holding the CL531  
face up in one hand, carefully separate the top and bottom of the housing. Place the unit into  
d
calibration mode by putting the switch into the OFF position and shorting the calibration via labele  
on the PCB with a shorting jumper or tweezers. Verify the CL531 is in calibration mode by  
viewing the LCD for the words “CALIBRATION REQUIRED”.  
Turn the CL531 on by moving the on/off switch to either “Read” or “Source”. Connect the  
CL531 terminals to a Current Source.  
1. Slide the EZ-Check™ Switch to the LO position indicated by displaying a “LO” on the  
left side of the LCD.  
2. Set the current source to 0.000 mA.  
3. Press and hold the EZ-Dial in for a few seconds.  
4. The display will flash “STORED” to confirm that the displayed calibration constant  
value was stored.  
1. Slide the EZ-Check™ Switch to the HI position indicated by displaying a “HI” on the  
left side of the LCD.  
2. Set the current source to 20.000 mA.  
3. Press and hold the EZ-Dial in for a few seconds.  
4. The display will flash “STORED” to confirm that the displayed calibration constant  
value was stored.  
CL  
Turn the unit off then back on to source mode by putting the on/off switch in the high position.  
Connect the Model 531 terminals to a Current Meter.  
1. Slide the EZ-Check™ Switch to the LO position indicated by displaying a “LO” on the  
left side of the LCD.  
2. Allow the current meter to settle and for the microprocessor to store 4mA.  
1. Slide the EZ-Check™ Switch to the HI position indicated by displaying a “HI” on the  
left side of the LCD.  
2. Allow the current meter to settle and for the microprocessor to store 20mA.  
Check the linearity calibration in source mode by dialing the Model 531 from 4.00mA to 4.01mA.  
If the linearity is calibrated correctly the unit will go from 4.00mA to 4.01mA. If this fails, repeat  
the calibration from the start.  
This is the completion of calibration.  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
NOTES  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARRANTY/DISCLAIMER  
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and  
workmanship for a period of 37 months from date of purchase. OMEGA’s WARRANTY adds  
an additional one (1) month grace period to the normal three (3) year product warranty to  
cover handling and shipping time. This ensures that OMEGA’s customers receive maximum  
coverage on each product.  
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer  
Service Department will issue an Authorized Return (AR) number immediately upon phone or  
written request. Upon examination by OMEGA, if the unit is found to be defective, it will be  
repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting  
from any action of the purchaser, including but not limited to mishandling, improper  
interfacing, operation outside of design limits, improper repair, or unauthorized modification.  
This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows  
evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture  
or vibration; improper specification; misapplication; misuse or other operating conditions  
outside of OMEGA’s control. Components in which wear is not warranted, include but are not  
limited to contact points, fuses, and triacs.  
OMEGA is pleased to offer suggestions on the use of its various products. However,  
OMEGA neither assumes responsibility for any omissions or errors nor assumes  
liability for any damages that result from the use of its products in accordance with  
information provided by OMEGA, either verbal or written. OMEGA warrants only  
that the parts manufactured by the company will be as specified and free of  
defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY  
KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL  
IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND  
FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF  
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total  
liability of OMEGA with respect to this order, whether based on contract, warranty,  
negligence, indemnification, strict liability or otherwise, shall not exceed the  
purchase price of the component upon which liability is based. In no event shall  
OMEGA be liable for consequential, incidental or special damages.  
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1)  
as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or  
activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or  
with any nuclear installation or activity, medical application, used on humans, or misused in  
any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER  
language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from  
any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.  
RETURN REQUESTS/INQUIRIES  
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department.  
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN  
AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT  
(IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be  
marked on the outside of the return package and on any correspondence.  
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to  
prevent breakage in transit.  
FOR WARRANTY RETURNS, please have  
the following information available BEFORE  
contacting OMEGA:  
1. Purchase Order number under which  
the product was PURCHASED,  
2. Model and serial number of the product  
under warranty, and  
3. Repair instructions and/or specific  
problems relative to the product.  
FOR NON-WARRANTY REPAIRS, consult  
OMEGA for current repair charges. Have the  
following information available BEFORE  
contacting OMEGA:  
1. Purchase Order number to cover the  
COST of the repair,  
2. Model and serial number of the  
product, and  
3. Repair instructions and/or specific problems  
relative to the product.  
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible.  
This affords our customers the latest in technology and engineering.  
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.  
© Copyright 2005 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,  
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without  
the prior written consent of OMEGA ENGINEERING, INC.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Where Do I Find Everything I Need for  
Process Measurement and Control?  
OMEGA…Of Course!  
Shop online at omega.com  
TEMPERATURE  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies  
Wire: Thermocouple, RTD & Thermistor  
Calibrators & Ice Point References  
Recorders, Controllers & Process Monitors  
Infrared Pyrometers  
PRESSURE, STRAIN AND FORCE  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Transducers & Strain Gages  
Load Cells & Pressure Gages  
Displacement Transducers  
Instrumentation & Accessories  
FLOW/LEVEL  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Rotameters, Gas Mass Flowmeters & Flow Computers  
Air Velocity Indicators  
Turbine/Paddlewheel Systems  
Totalizers & Batch Controllers  
pH/CONDUCTIVITY  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
pH Electrodes, Testers & Accessories  
Benchtop/Laboratory Meters  
Controllers, Calibrators, Simulators & Pumps  
Industrial pH & Conductivity Equipment  
DATA ACQUISITION  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Data Acquisition & Engineering Software  
Communications-Based Acquisition Systems  
Plug-in Cards for Apple, IBM & Compatibles  
Datalogging Systems  
Recorders, Printers & Plotters  
HEATERS  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Heating Cable  
Cartridge & Strip Heaters  
Immersion & Band Heaters  
Flexible Heaters  
Laboratory Heaters  
ENVIRONMENTAL  
MONITORING AND CONTROL  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Metering & Control Instrumentation  
Refractometers  
Pumps & Tubing  
Air, Soil & Water Monitors  
Industrial Water & Wastewater Treatment  
pH, Conductivity & Dissolved Oxygen Instruments  
M4207/0605  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Multi Tech Systems Network Card V34 336K User Manual
NetComm Network Router NP5400 User Manual
Niles Audio Car Amplifier SI 250 User Manual
Nilfisk ALTO Pressure Washer 5 SB FA User Manual
Nokia Cell Phone 1200 User Manual
Optoma Technology Projection Television ES520 User Manual
Oregon Scientific Clock CW101 User Manual
OWI Speaker SAT823 User Manual
Panasonic Cell Phone EB GD35 User Manual
Panasonic Network Card 2SA1309A User Manual