Oceanic Personal Computer Pro Plus 2 User Manual

OCEANIC  
®
Pro Plus 2  
dive computer  
operating manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CONTENTS  
LIMITED TWO-YEAR WARRANTY ................................................................................................................................ 2  
NOTICES ...................................................................................................................................................................... 2  
DECOMPRESSION MODEL ......................................................................................................................................... 2  
FEATURESAND DISPLAYS.............................................................................................. 7  
INTERACTIVE CONTROL CONSOLE .......................................................................................................................... 8  
BAR GRAPHS .............................................................................................................................................................. 9  
Tissue Loading Bar Graph (TLBG) ............................................................................................................................. 9  
OxygenBarGraph(O2BG) ...................................................................................................................................... 10  
VariableAscent Rate Indicator (VARI) ....................................................................................................................... 10  
AirTime Remaining Bar Graph (ATRBG) .................................................................................................................. 11  
DIVE TIME REMAINING .............................................................................................................................................. 11  
ALPHA/ NUMERIC DISPLAYS ................................................................................................................................... 12  
Cylinder Pressure ................................................................................................................................................... 12  
Depth Displays ....................................................................................................................................................... 12  
Time and Date Displays .......................................................................................................................................... 13  
Temperature Display ............................................................................................................................................... 13  
AUDIBLEALARM ....................................................................................................................................................... 14  
LED Warning Light .................................................................................................................................................. 14  
SMARTGLO BACKLIGHT ........................................................................................................................................... 16  
POWER SUPPLY ........................................................................................................................................................ 16  
Battery Consumption Indicator .................................................................................................................................. 16  
Low Battery Condition ............................................................................................................................................. 17  
FO2 MODE .................................................................................................................................................................. 18  
FO2 50% Default .................................................................................................................................................... 19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3
CONTENTS (continued)  
ACTIVATIONAND SETUP............................................................................................... 21  
ACTIVATION ............................................................................................................................................................... 22  
SURFACEMODE ........................................................................................................................................................ 23  
Accessing Other Modes .......................................................................................................................................... 23  
SET MODES ............................................................................................................................................................... 24  
ENTERING SETTINGS -SET MODE #1 ................................................................................................................... 25  
ENTERING SETTINGS -SET MODE #2 ................................................................................................................... 29  
PLANAND DIVE MODES ................................................................................................ 45  
TM  
PRE DIVE PLANNING SEQUENCE  
....................................................................................................................... 46  
TISSUE LOADING BAR GRAPH ................................................................................................................................. 48  
OXYGENACCUMULATION BAR GRAPH ................................................................................................................... 49  
VARIABLEASCENT RATE INDICATOR ..................................................................................................................... 49  
AIR TIME REMAINING BAR GRAPH BAR GRAPH ..................................................................................................... 50  
CONTROL OF DISPLAYS ........................................................................................................................................... 50  
NO DECOMPRESSION DIVE MODE .......................................................................................................................... 51  
DECOMPRESSION DIVE MODE ................................................................................................................................ 54  
VIOLATION MODES .................................................................................................................................................... 57  
Conditional Violation Mode ...................................................................................................................................... 57  
Delayed Violation Modes ......................................................................................................................................... 59  
Immediate Violation Mode and Gauge Mode ............................................................................................................ 60  
HIGH PO2 DIVE MODE ............................................................................................................................................... 62  
HIGH OXYGENACCUMULATION ............................................................................................................................... 63  
USER SET DIGITAL GAUGE MODE ........................................................................................................................... 64  
UNEXPECTED LOSS OF DISPLAYED INFORMATION .............................................................................................. 64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4
CONTENTS (continued)  
POST DIVE MODES ........................................................................................................ 65  
POSTDIVESURFACEMODE .................................................................................................................................... 66  
TRANSITION PERIOD ................................................................................................................................................ 66  
AFTER THE TRANSITION PERIOD (THE FIRST 2 HOURS) ....................................................................................... 68  
To activate the Backlight - ........................................................................................................................................ 68  
To access the Pre Dive Planning Sequence - ............................................................................................................ 68  
To access the Time to Fly Countdown ....................................................................................................................... 69  
To access the Time to Desaturate Countdown ........................................................................................................... 69  
Log Mode ............................................................................................................................................................... 70  
AFTER THE FIRST 2 HOURS ..................................................................................................................................... 72  
WET CONTACTS ........................................................................................................................................................ 73  
DOWNLOADING DATATOAPC ................................................................................................................................. 74  
SIMULATOR (DEMO) MODE ............................................................................................ 75  
GENERAL ........................................................................................................................ 83  
CAREAND CLEANING .............................................................................................................................................. 84  
INSPECTIONSAND SERVICE ................................................................................................................................... 84  
BATTERY REPLACEMENT ........................................................................................................................................ 86  
SPECIFICATIONS ....................................................................................................................................................... 90  
OCEANIC WORLD WIDE ............................................................................................................................................ 96  
SERVICE RECORD .................................................................................................................................................... 97  
RESET PROCEDURE ................................................................................................................................................ 99  
Payspecialattentiontoitemsmarked  
withthisWarningsymbol.  
WARNING: PriortodivingwiththeProPlus2,youmustalsoreadand  
understandtheOceanicDiveComputerSafetyandReferenceManual.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5
Components:  
a. Select(right)Button  
b. LEDWarningLight  
c. SmartgloSensingPort  
d. Advance(front)Button  
e. AirTimeRemaining BarGraph  
f. O2(oxygen)BarGraph  
g. Icon-DescendArrow  
Icon-DecompressionCeiling  
Icon-AscendArrow  
h. Graphic-MaximumDepth  
i. TissueLoadingBarGraph  
j. Battery Indicator  
k. Icon-Alarm(SetMode)  
l. Icon-LogMode  
m
n
l
o
DECO  
N
M
p
k
j
FEET  
MAX  
TOO  
FAS T  
q
DEMO  
i
MAX  
m. Graphic-DemoMode  
n. Graphic-Depth  
h
r
o
Graphic-AscentTooFast  
g
O2  
p. Icon-Time  
q. VariableAscentRateIndicator  
r. Icon-OperatingMode  
O2  
O2  
P SI  
BAR  
f
0
5
ATR  
1 0  
2 0  
3 0 4 0  
5 0 6 0  
e
a
d
b
c
PRO PLUS 2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6
WARNING: PriortodivingwiththeProPlus2,you  
mustalsoreadandunderstandtheOceanicDive  
ComputerSafetyandReferenceManualwhich  
providesImportantWarningsandSafetyRecom-  
mendationsaswellasgeneralproductinformation.  
FEATURES and DISPLAYS  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7
INTRODUCTION  
Welcome to Oceanic and thank you for choosing the Pro Plus 2 !  
It is extremely important that you read this Operating Manual in sequence and under-  
stand it completely before attempting to use the Pro Plus 2.  
It is equally important that you read the Oceanic Dive Computer Safety and Reference  
Manual (Doc. No. 12-2262) provided with your Pro Plus 2. It contains information that  
you must become familiar with prior to diving with your Pro Plus 2.  
Remember that technology is no substitute for common sense, and a dive computer only  
provides the person using it with data, not the knowledge to use it.  
INTERACTIVECONTROLCONSOLE  
DECO  
N
M
FEET  
MAX  
TOO  
FAS T  
The Interactive Control Console consists of two Control Buttons  
that allow you to select display options and access specific  
information when you want to see it. They are also used to  
enter Settings, activate the Backlight, and acknowledge the  
Audible Alarm.  
DEMO  
MAX  
O2  
O2  
O2  
0
P SI  
BAR  
5
1
0
2
0
3
0
4
0
5
0
6
0
ATR  
The Left button is referred to as Advance (Fig. 1a) and the  
Right button as Select (Fig. 1b).  
a
b
Fig.1-InteractiveControlConsole  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8
BARGRAPHS  
TissueLoadingBarGraph(TLBG)  
The Tissue Loading Bar Graph (Fig. 2a) represents tissue  
loading of nitrogen, showing your relative no decompression  
or decompression status. As your depth and elapsed dive  
time increase, segments will add to the Graph, and as you  
ascend to shallower depths, the Bar Graph will begin to  
recede, indicating that additional no decompression time is  
allowed for multilevel diving.  
The Tissue Loading Bar Graph monitors 12 different nitrogen  
compartments simultaneously and displays the one that is in  
control of your dive. It is divided into a green No Decom-  
pression (normal) zone, a yellow Caution zone (also No  
Decompression), and a red Decompression (danger) zone.  
DECO  
N2  
While you cannot provide a guarantee against the occur-  
rence of decompression sickness, you may choose your own  
personal zone of caution based upon age, physique, exces-  
sive weight, etc., to reduce the statistical risk.  
a
O2  
NOTE: DisplaysassociatedwithoxygenandtheO2  
BarGraphwillonlyappearifFO2hasbeensetata  
valueotherthan'Air'(e.g.,anumericalvalue).  
ATR  
Fig.2-TLBG  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9
OxygenAccumulationBarGraph(O2BG)  
The O2 Bar Graph (Fig. 3a) represents oxygen loading, showing the maximum of either  
per dive accumulated oxygen, or 24 hour period accumulated oxygen.  
As your oxygen exposure (accumulation) increases during the dive, segments will add to  
the Bar Graph, and as loading decreases, it will begin to recede, indicating that addi-  
tional exposure is allowed for that dive and 24 hour period.  
VariableAscentRateIndicator(VARI)  
The Variable Ascent Rate Indicator (Fig. 3b) provides a visual representation of ascent  
speed (i.e., an ascent speedometer). Green is a 'normal' rate, yellow a 'caution' rate,  
and red is 'Too Fast'. The segments of the VARI represent two sets of speeds which  
change at a reference depth of 60 feet (18 meters). Refer to the chart (Fig. 4) for seg-  
ment values.  
60 feet (18 m) & Shallower  
Deeper than 60 feet (18 m)  
Segments  
Ascent Rate =  
Segments  
Ascent Rate =  
DECO  
N2  
Displayed  
FPM  
0-10  
11-15 3.5-4.5  
16-20 5-6  
21-25 6.5-7.5  
26-30 8-9  
>30  
MPM  
0 - 3  
Displayed  
FPM  
0-20  
21-30 6.5-9  
31-40 9.5-12  
41-50 12.5-15  
51-60 15.5-18  
MPM  
0 - 6  
b
0
1
2
3
4
5
0
1
2
3
4
5
TOO  
FAS T  
>9  
>60  
>18  
a
Fig.4-VARIValues  
WARNING: Atdepthsgreaterthan60feet(18  
meters),ascentratesshouldnotexceed60fpm(18  
mpm). Atdepthsof60feet(18meters)andshallower,  
ascentratesshouldnotexceed30fpm(9mpm).  
O2  
O2  
0
5
1 0  
2 0  
3 0 4 0  
5 0 6 0  
ATR  
Fig.3-O2BGandVARI  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10  
N2  
AirTimeRemainingBarGraph(ATRBG)  
FEET  
The Air Time Remaining Bar Graph (Fig. 5a) provides a  
graphic representation of the time that you can remain at your  
present depth and then, following a safe ascent, surface with a  
predetermined reserve of breathing gas.  
This calculation and display is based on your Breathing Gas  
Consumption Rate that is continuously monitored by the Pro  
Plus 2, and it takes into account the breathing gas required for  
a safe ascent including any required decompression stops.  
b
O2  
P SI  
a
0
5
ATR  
1 0  
2 0  
Fig.5-ATRBG  
The green, yellow, and red zones adjacent to the bar graph  
enable you to quickly focus on remaining breathing gas times of 60 minutes and less,  
based on your pre selected End Pressure Alarm Set Point (described later). The bar  
graph is more precise as time decreases toward the red zone.  
DIVETIMEREMAINING  
The Pro Plus 2 constantly monitors three critical pieces of information; no decompression  
status, oxygen accumulation, and breathing gas consumption rate. Refer to the Oceanic  
Dive Computer Safety and Reference Manual which describes Dive Time Remaining  
which is a display that indicates the time that is more critical for you at that particular  
moment (i.e.; whichever time is the least amount available of the three).  
Displayed will be either Time Remaining before reaching the No Decompression limit  
(Fig. 5b), or Time Remaining before reaching the limit for Oxygen Accumulation, or Time  
Remaining before reaching the End Pressure Alarm set point.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
11  
ALPHA/NUMERICDISPLAYS  
CylinderPressureDisplay  
Cylinder Pressure (Fig 6a) is displayed any time the Pro Plus  
2 is connected to a pressurized cylinder and in an active  
operating mode.  
Values of pressure are displayed numerically from 10 PSI (.5  
BAR) up to 5000 PSI (352 BAR) in increments of 10 PSI (.5  
BAR).  
DepthDisplays  
During a dive, the Current Depth display (Fig. 6b), indi-  
cates depths from 0 to 330 feet (99.9 meters) in 1 foot (.1  
meter) increments.  
b
N2  
FEET  
By pressing the Left (Advance) button, the Maximum  
Depth reached during that dive will be displayed in the  
center/left portion of the display (Fig. 6c).  
MAX  
c
During a Decompression Dive, the required Ceiling Stop  
Depth is displayed in the center of the screen. Maximum  
Depth can be viewed by pressing the Left (Advance) button.  
O2  
P SI  
0
5
1 0  
2 0  
ATR  
a
Fig.6-CylinderPressure  
andDepth Displays  
Download from Www.Somanuals.com. All Manuals Search And Download.  
12  
TimeandDateDisplays  
Time displays are shown in hour:minute format (i.e., 1:16 represents 1 hour and 16  
minutes, not 116 minutes!). The colon that separates hours and minutes blinks once per  
second when the display is indicating real time (e.g., Elapsed Dive Time), and is solid  
(non-blinking) when times are calculated projections (e.g., Time to Fly).  
The Main Time display is located in the lower/middle portion of the display (Fig. 7a)  
and a second time display (Fig. 7b) is located in the center/right. Both displays are  
identified by clock icons.  
• Time of Day can be set for 12 hour format (Am/Pm) or 24 hour format.  
Date is displayed in the center/left portion of the screen only  
to identify dive data while it is viewed in the Log Mode (see  
b
page 70). When Units of Measure are set for 'Imperial', the  
N2  
FEET  
Month appears to the left of Day. When set for Metric, the  
Month appears to the right of Day.  
TemperatureDisplay  
Ambient Temperature is displayed in the center/left  
portion of the screen (Fig. 7c) while in the Surface Mode and  
c
Log Mode, and can be viewed as part of an Alternate Display  
when the Left (Advance) button is pressed while in a dive  
mode. If the Temperature exceeds a value of '99', two dashes  
( - - ) will be displayed on the screen until the unit's tempera-  
ture decreases to '99'.  
a
O2  
P SI  
0
5
ATR  
1 0  
2 0  
Fig.7-TimeandTemperature  
Displays  
Download from Www.Somanuals.com. All Manuals Search And Download.  
13  
NOTE: Eachnumericandgraphicdisplayrepresentsauniquepieceofinforma-  
tion. Itisimperativethatyouunderstandtheformats,ranges,andvaluesofthe  
informationrepresentedtoavoidanypossiblemisunderstandingthatcouldresult  
inerror. TheInformationalDisplaysaredescribedindetailasthevariousoperat-  
ingmodestheyappearinarepresentedthroughoutthismanual.  
AUDIBLEALARM  
LEDWarningLight  
A red LED Warning Light will light and a Speaker icon will appear when the Audible  
Alarm emits a tone. They will turn off when the Alarm is acknowledged, and will not  
activate if the Audible is Set OFF (a user setting).  
When warning situations activate the Alarm, the unit will emit a continuous tone for 10  
seconds, or until the situation is corrected, or it is acknowledged by the user pressing the  
Left (Advance) button for 2 seconds. If acknowledged by the user and the situation  
corrected, the Alarm will sound again upon reentry into the warning situation, or entry  
into another type of warning situation.  
Asingleshortbeep(whichcannotbedisabled)isemitted forthefollowing-  
• After the Diagnostic countdown, if everything is okay.  
• If the unit automatically returns to Surface Mode from Simulator Mode.  
• Upon completion of a Hot Swap battery change with calculations saved.  
• Change from Delayed to Full Violation 5 minutes after the dive.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
14  
SituationsthatwillsoundtheAlarm,ifit isturnedON(ausersetting),include-  
• Entry into Decompression Mode.  
• Air Time Remaining Bar Graph Alarm (a user setting).  
• Air Time Remaining = 5 minutes.  
• Air Time Remaining = 0 minutes.  
• Turn Pressure Alarm (a user setting).  
• End Pressure Alarm (a user setting).  
• PO2 => than the Max PO2 Alarm (a user setting), or => 1.60 ATA.  
• Descent deeper than the Max Depth Alarm (a user setting).  
• Tissue Loading Bar Graph Alarm (a user setting).  
• Dive Time Remaining Alarm (a use setting).  
• Elapsed Dive Time Alarm (a use setting).  
• O2 Accumulation => allowable per dive limit, or limit for a 24 hour period.  
• Ascending above a required Decompression ceiling stop depth for less than 5 minutes  
(referred to as a Conditional Violation).  
• Ascent rate exceeds 60 feet/minute (18 meters per minute) if greater than 60 feet (18  
meters), or 30 feet/minute (9 meters/minute) at 60 feet (18 meters) and shallower.  
Duringthefollowingsituations,the10secondcontinuoustonewillbefollowedbya5  
secondsteadybeepthatwillnotturnoffwhenacknowledged,evenifitwasuserSet  
OFF-  
• Ascending above a required Decompression ceiling stop depth for more than 5  
minutes (referred to as a Delayed Violation).  
• Decompression requires a ceiling stop depth of 70 feet/21 meters or deeper.  
• Being on the surface for 5 minutes after a Conditional Violation (Permanent Violation).  
Download from Www.Somanuals.com. All Manuals Search And Download.  
15  
SMARTGLO®BACKLIGHT  
To activate the Backlight - press the Right (Select) button.  
• The Smart Glo® feature senses the intensity of natural  
light present. If a low level of light is present, the Back-  
light will activate and illuminate the display for button  
depression time plus the user set additional Duration time  
(3 or 7 seconds).  
• Press the button again to activate as desired.  
NOTE: ExtensiveuseoftheBacklightreduces  
estimatedBatterylife. Also,theBacklightdoesnot  
operateduringaLowBatteryConditionorwhen  
downloadingdatatoaPC.  
a
POWERSUPPLY  
N2  
The Pro Plus 2 utilizes one (1) 3 volt, CR2 Lithium Battery that  
should provide from 50 dive hours of operation, if you  
conduct 1 - one hour dive each time the unit is activated, to  
over 150 dive hours of operation, if you conduct 3 or more  
one dives each time the unit is activated.  
Battery Indicator  
A Battery Indicator (Fig. 8a) provides an indication of Battery  
condition. The Battery Indicator will be displayed during  
Surface Mode. It will not be displayed during Dive Mode(s).  
O2  
ATR  
Fig.8-Battery Indicator  
Download from Www.Somanuals.com. All Manuals Search And Download.  
16  
LowBatteryCondition  
• Voltage level is checked upon activation and every 10  
minutes during operation.  
• When 75 % of the Rated Power has been consumed,  
only the lower bar of the Indicator will be displayed and  
the full icon will flash (Fig. 9) once per second as a  
warning that the Battery is to be replaced prior to con-  
ducting any further dives.  
• Upon decreasing to a voltage level that will no longer  
sustain proper operation, the Battery Indicator will flash 5  
times followed by shutdown of the unit.  
• If a Low Battery Condition exists when the unit is manu-  
ally activated (by pressing the button), the graphic bAT  
and the Battery Indicator will appear flashing for 5  
seconds followed by shutdown of the unit.  
N2  
• If the button is not pressed to activate the unit prior to a  
dive (e.g., the unit activated automatically by immersion  
in water), and a Low Battery Condition exists, the Battery  
Indicator will appear flashing as a warning upon descent  
past 4 feet (1.2 meters). No other information will be  
displayed and the unit will not enter Dive Mode.  
O2  
P SI  
ATR  
Fig.9-LowBatteryCondition  
Download from Www.Somanuals.com. All Manuals Search And Download.  
17  
• If the unit did not display a Low Battery Condition prior to  
entering the Dive Mode, and a Low Battery Condition  
occurs during the dive, there will be sufficient Battery  
power to maintain unit operation for the remainder of  
that dive. The Battery Indicator will appear after the dive  
upon entry into Surface Mode.  
• When the Battery is removed, settings and calculations  
for repetitive dives will be retained, if a new battery can  
be inserted within 8 seconds. Otherwise the calculations  
will reset to zero and settings must be reset.  
FO2MODE  
After Activation, the Pro Plus 2 will operate as an Air com-  
puter without displaying information associated with oxygen  
calculations, unless it is set for a percentage of oxygen (FO2)  
other than Air (a numerical value between 21 and 50 %).  
N2  
When set with an FO2 value of 'Air' (Fig. 10), the Pro  
Plus 2 will perform calculations the same as if FO2 were set  
for 21% oxygen, internally accounting for oxygen loading for  
any subsequent Nitrox dives. However, oxygen related  
displays, warnings, and the O2 Bar Graph will not appear  
on the display for that dive, or subsequent dives, unless FO2  
is set for a numerical value (21 - 50).  
O2  
ATR  
Fig.10-FO2setforAIR  
Download from Www.Somanuals.com. All Manuals Search And Download.  
18  
N2  
Once a dive is made with the unit set as a Nitrox Computer  
(FO2 set for a numerical value), it cannot be programmed to  
operate as an Air Computer until 24 hours after the last dive.  
'Air' will not be displayed as an option in the FO2 Mode.  
However, you can set FO2 for 21% for use with Air.  
MAX  
When FO2 is set at a value of 21% (Fig. 11), the unit will  
remain set at 21% for subsequent nitrox dives until FO2 is  
set to a higher value, or until it automatically turns off and is  
reactivated.  
O2  
ATR  
Fig.11-FO2setfor21%  
FO250%Default  
If the Default function is set to ON (Fig. 12) and FO2  
is set to a value 'greater than 21%', the FO2 set point value  
will automatically revert to 50% 10 minutes after that dive.  
The Maximum Depth that can be achieved with a PO2 of  
1.60 ATA will also be displayed.  
N2  
• FO2 must therefore be reset for each repetitive nitrox  
dive, or the value will automatically 'default' to 50(%)  
and the dives will be calculated based on 50% O2 (50%  
nitrogen) for oxygen calculations and 21% O2 (79%  
nitrogen) for nitrogen calculations.  
O2  
ATR  
Fig.12-FO2DefaultON  
Download from Www.Somanuals.com. All Manuals Search And Download.  
19  
If the Default function is set to OFF (Fig. 13), the FO2  
value for repetitive dives remains the same as previously set  
until the set point is manually changed.  
WARNING: Whentheunitshutsoff,theFO2setpoint  
willdefaulttoAIRregardlessifthe50%Defaultisset  
ONorOFF. Alwaysset, orverify, theFO2setpoint  
priortoeachnitroxdive.  
SettingFO2foranitroxdiveisdescribedonpage25.  
N2  
O2  
ATR  
Fig.13-FO2DefaultOFF  
Download from Www.Somanuals.com. All Manuals Search And Download.  
20  
WARNING: PriortodivingwiththeProPlus2,you  
mustalsoreadandunderstandtheOceanicDive  
ComputerSafetyandReferenceManualwhich  
providesImportantWarningsandSafetyRecom-  
mendationsaswellasgeneralproductinformation.  
ACTIVATION and SETUP  
Download from Www.Somanuals.com. All Manuals Search And Download.  
21  
ACTIVATION  
To Activate the Pro Plus 2, press and release the Left (Advance) Button.  
• Upon manual activation, the unit will enter Diagnostic Mode (Fig. 14), displaying all  
segments of the LCD as 8's, followed by dashes (- -), then a countdown from 9 to 0.  
Diagnostic Mode checks the display and battery voltage to ensure that everything is  
within tolerance and functioning properly.  
BackupActivation(onlyifWaterActivationissetON)  
As a backup, the Pro Plus 2 will also automatically activate by water contact. This is  
accomplished by bridging the gap between contacts located on the Button stems and  
case. The graphic H2O that will be displayed as an indication is described later.  
If no dive is made within 2 hours after initial activation, the unit  
DECO  
N2  
M
FEET  
MAX  
will automatically deactivate. If the wet contacts are still  
TOO  
bridged, the unit will reactivate and display the H2O graphic.  
FAS T  
DEMO  
WARNING: Iftheunitismanuallyactivatedatelevations  
higherthan14,000feet(4,267meters),itwillperforma  
diagnosticcheckfollowedbyimmediateshutdown.  
MAX  
O2  
O2  
O2  
P SI  
BAR  
0
5
1 0  
2 0  
3 0 4 0  
5 0 6 0  
ATR  
Fig.14-DiagnosticMode  
Download from Www.Somanuals.com. All Manuals Search And Download.  
22  
SURFACEMODE  
N2  
Surface Mode (Fig. 15), identified by the Surface Time icon,  
follows Diagnostic Mode after Activation. Information  
includes Dive Number '0' (no dive made yet), Temperature  
(and icon), Time of Day (with icon), the Battery Consumption  
Indicator, and Surface Time (with flashing colon).  
NOTE: Ifthewetcontactsarebridged,thegraphic  
'H2O'willappearinplaceofthedivenumber'0'(Fig.  
16). Aftertheunitisrinsedanddried,'0'willreplace  
'H2O'.  
O2  
P SI  
ATR  
Fig.15-SurfaceMode  
To activate the Backlight:  
• press the Right (Select) button.  
N2  
AccessingOtherModes  
While in the Surface Mode you can access various other  
Modes that are described in detail throughout this manual.  
• Press the Left (Advance) button to access a sequence that  
includes Plan, Fly, DeSaturation, and Log Modes.  
• Press the Right (Select) button to activate the Backlight.  
• Press Both buttons (simultaneously) to access Set and  
Simulator Modes.  
O2  
P SI  
ATR  
Fig.16-SurfaceMode  
(rinseanddrytheunit)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
23  
SETMODES  
N2  
Settings are divided into 2 categories. Set Mode #1 includes  
several settings that you would change more often and Set  
Mode #2 includes those items not likely to change once you  
set them. Set Mode 2 can be accessed by first entering  
settings in Set Mode 1, or by bypassing Set Mode 1.  
O2  
Settings can be made one after the other, or you can access  
a specific item that you want to set, bypassing others. Set  
points can be advanced by scrolling (press/hold) or one  
increment at a time (press/release < 2 seconds)  
ATR  
Fig.17-SetMode1  
Set Mode Access Timing  
While in Surface Mode, press Both buttons simultaneously  
and hold -  
N2  
• after 2 seconds, SET: 1 appears (Fig. 17)  
• after 2 more seconds, SET: 2 appears (Fig. 18)  
• Access is gained by releasing the buttons during the 2  
second window in which SET: 1 or SET: 2 appears. then  
pressing the Left (Advance) button.  
• If the buttons are held longer, and SET 1 and 2 are both  
bypassed, the unit will go to Simulator (Demo) Mode  
which is described on page 75.  
O2  
• While in the Set Mode, if neither button is pressed during  
a 2 minute period, the unit will revert to Surface Mode.  
ATR  
Fig.18-SetMode2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
24  
N2  
ENTERING SETTINGS -SET MODE #1  
TO SET - FO2 (while in the Surface Mode)  
Factory set for AIR, FO2 can also be set to values between  
21 and 50% in increments of 1%.  
• Press Both buttons simultaneously, release when SET: 1  
appears (2 seconds).  
O2  
• Press and release the Left (Advance) button, FO2 ap-  
pears with the set point value flashing (Fig. 19).  
ATR  
Fig.19-SetFO2  
• Press and release the Right (Select) button repeatedly to  
increase the FO2 value from 21 to 50% in increments of  
1%, then display AIR again; - or - Press and hold the  
Right (Select) button to scroll from AIR to 32%, then press  
and hold again to scroll from 32 to 50%, then AIR.  
N2  
• For each FO2 value that appears, the display indicates  
the Maximum Depth that can be achieved for a PO2 of  
1.60 ATA (Fig. 20a), or the PO2 Alarm set point. If FO2  
is set for AIR, no Depth value is displayed.  
MAX  
a
O2  
• Press the Left (Advance) button to accept the setting and  
advance to Set Depth Alarm, or press and hold Both  
buttons for 2 seconds to revert to Surface Mode.  
ATR  
Fig.20-FO2setfor32%  
Download from Www.Somanuals.com. All Manuals Search And Download.  
25  
TO SET - MAX DEPTH ALARM  
(while in the Surface Mode)  
Factory set for 330 feet, the Alarm can be set to values  
between 30 feet (3 meters) and 330 feet ( 99 meters) in  
increments of 10 feet (3 meters).  
• Press Both buttons simultaneously, release when SET: 1  
appears (2 seconds).  
• Press and release the Left (Advance) button, FO2 ap-  
pears with the value flashing.  
• Press the Left (Advance) button 1 more time.  
N2  
FEET  
MAX  
• The graphics FEET MAX and dEEP, and Alarm icon  
appear with the Max Depth set point value flashing  
(Fig. 21).  
• Press and release the Right (Select) button until the  
desired Depth Alarm value appears; or press and hold to  
scroll through the set points.  
O2  
• Press the Left (Advance) button to accept the setting and  
advance to Elapsed Dive Time Alarm, or press and hold  
Both buttons for 2 seconds to revert to Surface Mode.  
ATR  
Fig.21-SetMaxDepthAlarm  
Download from Www.Somanuals.com. All Manuals Search And Download.  
26  
TO SET - ELAPSED DIVE TIME ALARM  
(while in the Surface Mode)  
Factory set for 3:00 (hr:min), the Alarm can be set to values  
between 0:10 and 3:00 (hr:min) in increments of 5 minutes.  
• Press Both buttons simultaneously, release when SET: 1  
appears (2 seconds).  
• Press and release the Left (Advance) button, FO2 ap-  
pears with the set point value flashing.  
• Press the Left (Advance) button 2 more times.  
• The graphic EdT, and Alarm and Dive Time icons appear  
with the Elapsed Dive Time value flashing (Fig. 22).  
N2  
• Press and release the Right (Select) button until the  
desired Alarm value appears; or press and hold to scroll  
through the set points.  
• Press the Left (Advance) button to accept the setting and  
advance to PC Interface, or press and hold Both buttons  
for 2 seconds to revert to Surface Mode.  
O2  
ATR  
Fig.22-SetElapsedDive  
TimeAlarm  
Download from Www.Somanuals.com. All Manuals Search And Download.  
27  
PC INTERFACE  
NOTE: For more  
information  
PC Interface is not a setting, it is included in the Set 1 menu  
for easy access when data in the unit's memory is to be  
downloaded (copied) to the PC download software program  
for storage and viewing.  
regardingPC  
Interface,referto  
page74ofthis  
manualandto  
documents  
To download data (while in the Surface Mode)-  
providedwiththe  
PCdownload  
product.  
• Press Both buttons simultaneously, release when SET: 1  
appears (2 seconds).  
• Press and release the Left (Advance) button, FO2 ap-  
pears with the set point value flashing.  
• Press the Left (Advance) button 3 more times.  
N2  
• The graphic PC appears with a 120 second countdown  
timer (Fig. 23). Download must be initiated before the  
timer reaches 00 (within 2 minutes).  
• Download is initiated by the external device requesting  
data transfer (i.e., the PC download program).  
O2  
• The unit reverts to Surface Mode after completion of the  
Download operation, or after 2 minutes if neither button  
is pressed.  
ATR  
Fig.23-PCInterface  
Download from Www.Somanuals.com. All Manuals Search And Download.  
28  
ENTERING SETTINGS -SET MODE #2  
To return to Sur-  
face Mode at any  
time while in Set  
Mode, press and  
hold Both buttons  
for 2 seconds.  
These settings are ones that are not likely to change. To save  
time at the dive site, verify the set points and adjust them as  
desired prior to departing on the day's dive trip.  
TO SET - UNITS OF MEASURE  
(while in the Surface Mode)  
Factory set for Imperial, Units of Measure can also be set for  
Metric.  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
N2  
• Press and release the Left (Advance) button, the Units  
screen appears with the graphics FT, F, and PSI (or M, C,  
and BAR) flashing (Fig. 24).  
• Press and release the Right (Select) button to toggle  
between Imperial and Metric units.  
• Press the Left (Advance) button to accept the setting and  
advance to Set Hour Format, or press and hold Both  
buttons for 2 seconds to revert to Surface Mode.  
O2  
ATR  
Fig.24-SetUnitsofMeasure  
Download from Www.Somanuals.com. All Manuals Search And Download.  
29  
TO SET - HOUR FORMAT (while in the Surface Mode)  
Factory set for 12 Hr (12: Am to 11: Pm), the Format can  
also be set for 24 Hr (0: to 23: hours).  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 1 more time.  
• The graphic Hour appears with 12 (or 24) flashing  
(Fig. 25).  
N2  
• Press and release the Right (Select) button to toggle  
between 12 and 24.  
• Press the Left (Advance) button to accept the setting and  
advance to Set Hour, or press and hold Both buttons for  
2 seconds to revert to Surface Mode.  
O2  
ATR  
Fig.25-SetHourFormat  
Download from Www.Somanuals.com. All Manuals Search And Download.  
30  
N2  
TO SET - TIME (while in the Surface Mode)  
Set for factory local time, the Time can be set to values  
between 1:00 and 12:59 (AM/PM) or 0:00 and 23:59.  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
O2  
ATR  
• Press the Left (Advance) button 2 more times.  
Fig.26-SetHour  
• Time appears with the Hour value flashing (Fig. 26).  
• Press and hold the Right (Select) button to scroll through  
the Hour values in increments of one hour.  
N2  
• Press the Left (Advance) button to accept the setting.  
• The Minute value flashes (Fig. 27).  
• Press and hold the Right (Select) button to scroll through  
the Minute values in increments of one minute.  
O2  
ATR  
• Press the Left (Advance) button to accept the setting and  
advance to Set Year.  
Fig.27-SetMinute  
Download from Www.Somanuals.com. All Manuals Search And Download.  
31  
N2  
TO SET - DATE (after setting Time)  
Factory set for the factory local Date, the Date can be set to  
values between 01/01/01 and 12/31/20.  
• After having set and accepted the Time, the Date ap-  
pears with the graphic dAY, and Year value flashing  
(Fig. 28).  
Fig.28-SetYear  
• Press and hold the Right (Select) button to scroll through  
the Year values in increments of one Year.  
N2  
• Press the Left (Advance) button to accept the setting.  
• The Month value flashes (Fig. 29).  
• Press and release the Right (Select) button to scroll  
through the Month values in increments of one month.  
Fig.29-SetMonth  
• Press the Left (Advance) button to accept the setting.  
N2  
• The Day value flashes (Fig. 30).  
• Press and hold the Right (Select) button to scroll through  
the Day values in increments of one day.  
• Press the Left (Advance) button to accept the setting.  
Fig.30-SetDay  
Download from Www.Somanuals.com. All Manuals Search And Download.  
32  
TO SET - AUDIBLE ALARM (while in the Surface Mode)  
Factory set for ON, the Alarm can be also be set to OFF.  
This setting also applies to the red LED Warning Indicator  
that is synchronized with the Audible Alarm.  
When set OFF, the Alarm will not sound during the condi-  
tions described on page 14.  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 7 more times.  
N2  
• The graphic ALM and Alarm (speaker) icon appear with  
On (or OFF) flashing (Fig. 31).  
• Press and release the Right (Select) button to toggle  
between ON and OFF.  
• Press the Left (Advance) button to accept the setting and  
advance to Set Max TLBG Alarm, or press and hold Both  
buttons for 2 seconds to revert to Surface Mode.  
O2  
ATR  
Fig.31-SetAudibleAlarm  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33  
TO SET - MAXIMUM TISSUE LOADING BAR GRAPH  
(TLBG) ALARM (while in the Surface Mode)  
Factory set for DECO (all 8 segments), the Maximum Tissue  
Loading Bar Graph (TLBG) Alarm can be set to values  
between DECO (all 12 segments) and 1 segment.  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 8 more times.  
DECO  
N2  
• The graphic ndc and Alarm (speaker) icon appear with  
the full TLBG flashing (Fig. 32).  
• Press and release the Right (Select) button to decrease  
the number of segments one at a time.  
• Press the Left (Advance) button to accept the setting and  
advance to Set Air Time Remaining Alarm, or press and  
hold Both buttons for 2 seconds to revert to Surface  
Mode.  
O2  
ATR  
Fig.32-SetMaxTLBGAlarm  
Download from Www.Somanuals.com. All Manuals Search And Download.  
34  
TO SET - AIR TIME REMAINING BAR GRAPH ALARM  
(while in the Surface Mode)  
Factory set for 5 minutes, the Air Time Remaining Bar Graph  
Alarm can be set to values of 0, 5, and 10 to 60 minutes in  
increments of 5 minutes.  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 9 more times.  
N2  
• The graphic Atr, and Alarm (speaker) and Cylinder icons  
appear with the Air Time Remaining Bar Graph  
value flashing (Fig. 33).  
• Press and release the Right (Select) button to decrease  
the Alarm value in increments of 5 minutes. Or press  
and hold to scroll.  
O2  
• Press the Left (Advance) button to accept the setting and  
advance to Set Turn Pressure Alarm, or press and hold  
Both buttons for 2 seconds to revert to Surface Mode.  
0
5
ATR  
1 0  
2 0  
3 0 4 0  
5 0 6 0  
Fig.33-SetAirTimeRemaining  
BarGraph Alarm  
Download from Www.Somanuals.com. All Manuals Search And Download.  
35  
TO SET - TURN PRESSURE ALARM  
(while in the Surface Mode)  
Factory set for 00 PSI/BAR (disabled), the Turn Pressure  
Alarm can be set to values between 1000 and 3000 PSI (69  
to 205 BAR) in increments of 250 PSI (17 BAR).  
• Press Both buttons simultaneously, release when SET: 2  
appears.  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 10 more times.  
N2  
• The graphic trn and Alarm (speaker) icon appear with  
the Turn Pressure Alarm value flashing (Fig. 34).  
• Press and hold the Right (Select) button to scroll through  
the Alarm values in increments of 250 PSI (17 BAR).  
• Press the Left (Advance) button to accept the setting and  
advance to Set End Pressure Alarm, or press and hold  
Both buttons for 2 seconds to revert to Surface Mode.  
O2  
P SI  
ATR  
Fig.34-SetTurnPressureAlarm  
Download from Www.Somanuals.com. All Manuals Search And Download.  
36  
TO SET - END PRESSURE ALARM  
(while in the Surface Mode)  
Factory set for 00 PSI/BAR (disabled), the End Pressure  
Alarm can be set to values between 300 and 1500 PSI (20  
to 104 BAR) in increments of 100 PSI (7 BAR).  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 11 more times.  
N2  
• The graphic End and Alarm (speaker) icon appear with  
the End Pressure Alarm value flashing (Fig. 35).  
• Press and hold the Right (Select) button to scroll through  
the Alarm values in increments of 100 PSI (6 BAR).  
• Press the Left (Advance) button to accept the setting and  
advance to Set Dive Time Remaining Alarm, or press and  
hold Both buttons for 2 seconds to revert to Surface  
Mode.  
O2  
P SI  
ATR  
Fig.35-SetEndPressureAlarm  
Download from Www.Somanuals.com. All Manuals Search And Download.  
37  
TO SET - DIVE TIME REMAINING ALARM  
(while in the Surface Mode)  
Factory set for 5 minutes, the Dive Time Remaining Alarm  
can be set to values between 0:00 and 0:20 (minutes) in  
increments of 1 minute.  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 12 more times.  
N2  
• The graphic dtr, and Alarm (speaker) and Dive Mode  
icons appear with the Dive Time Remaining value  
flashing (Fig. 36).  
• Press and hold the Right (Select) button to scroll through  
the Alarm values in increments of one minute.  
O2  
• Press the Left (Advance) button to accept the setting and  
advance to Set Max PO2 Alarm, or press and hold Both  
buttons for 2 seconds to revert to Surface Mode.  
ATR  
Fig.36-SetDiveTime  
RemainingAlarm  
Download from Www.Somanuals.com. All Manuals Search And Download.  
38  
TO SET - MAXIMUM PO2 ALARM  
(while in the Surface Mode)  
Factory set for 1.60 (ATA), the Maximum PO2 Alarm can be  
set to values between 1.20 and 1.60 (ATA) in increments of  
.10 (ATA).  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 13 more times.  
N2  
• The graphics PO2 and MAX, and Alarm (speaker) icon  
appear with the PO2 value flashing (Fig. 37).  
• Press and release the Right (Select) button to advance the  
Alarm value in increments of .10 (ATA).  
MAX  
• Press the Left (Advance) button to accept the setting and  
advance to Set FO2 50% Default, or press and hold Both  
buttons for 2 seconds to revert to Surface Mode.  
O2  
ATR  
Fig.37-MaximumPO2Alarm  
Download from Www.Somanuals.com. All Manuals Search And Download.  
39  
TO SET - FO2 50% DEFAULT  
(while in the Surface Mode)  
Factory set ON, the FO2 50% Default feature can be set to  
OFF. The effects of this feature being ON or OFF are  
described on page 19.  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 14 more times.  
N2  
• The graphics FO2 and 50 appear with ON (or OFF)  
flashing (Fig. 38).  
• Press and release the Right (Select) button to toggle  
between ON and OFF.  
• Press the Left (Advance) button to accept the setting and  
advance to Set Backlight Duration, or press and hold  
Both buttons for 2 seconds to revert to Surface Mode.  
O2  
ATR  
Fig.38-SetFO250%Default  
Download from Www.Somanuals.com. All Manuals Search And Download.  
40  
TO SET - BACKLIGHT DURATION  
(while in the Surface Mode)  
Factory set for 3 seconds, the Backlight Duration can be set  
to values of 0 (disabled), 3, or 7 seconds. This is the length  
of time the Backlight stays On after the button is released.  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 15 more times.  
N2  
• The graphic GLO and Time icon appear with the Dura-  
tion value flashing (Fig. 39).  
• Press and release the Right (Select) button to advance the  
Duration from :00 to :03 to :07 (seconds).  
• Press the Left (Advance) button to accept the setting and  
advance to Set Sampling Rate, or press and hold Both  
buttons for 2 seconds to revert to Surface Mode.  
O2  
ATR  
Fig.39-SetBacklightDuration  
Download from Www.Somanuals.com. All Manuals Search And Download.  
41  
TO SET - PC SAMPLING RATE  
(while in the Surface Mode)  
NOTE: Sampling  
Rateisthe  
intervalatwhich  
datasamplesare  
recordedduringa  
diveforsubse-  
quentdownload  
tothePCpro-  
gram. This  
Factory set at 15 seconds, the Sampling Rate can be set at  
values of 2, 15, 30, or 60 seconds, or 2, 5, or 10 feet (.5,  
1.5, or 3 meters).  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
settinghasno  
effectondis-  
playeddataor  
dataintheunit's  
viewableLog.  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 16 more times.  
N2  
• The graphics SR and SECS appear with the Sampling  
Rate value flashing (Fig. 40).  
• Press and release the Right (Select) button advance the  
Rate one selection at a time.  
• Press the Left (Advance) button to accept the setting and  
advance to Set Digital Gauge Mode, or press and hold  
Both buttons for 2 seconds to revert to Surface Mode.  
O2  
ATR  
Fig.40-SetPCSamplingRate  
Download from Www.Somanuals.com. All Manuals Search And Download.  
42  
TO SET - USER SET DIGITAL GAUGE MODE  
NOTE: Once a  
diveismadewith  
thisfeatureset  
ON,thesettingwill  
belockedONfor  
24hoursafterthe  
dive. SetDigital  
GaugeModewill  
notappearasa  
selectionduring  
that24hour  
(while in the Surface Mode)  
Factory set OFF, Digital Gauge Mode can also be set ON.  
The effects of this feature being ON or OFF are described on  
page 64.  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
period.  
• Press the Left (Advance) button 17 more times.  
N2  
• The graphic GAU appears with OFF (or ON) flashing  
(Fig. 41).  
• Press and release the Right (Select) button to toggle  
between ON and OFF.  
• Press the Left (Advance) button to accept the setting and  
advance to Set Water Activation, or press and hold Both  
buttons for 2 seconds to revert to Surface Mode.  
O2  
ATR  
Fig.41-SetDigitalGaugeMode  
Download from Www.Somanuals.com. All Manuals Search And Download.  
43  
TO SET - WATER ACTIVATION  
WARNING: Ifthis  
featureissetfor  
OFF,theProPlus  
2mustbemanu-  
allyactivatedprior  
tocommencinga  
dive.  
(while in the Surface Mode)  
Factory set for ON, the Water Activation feature can also  
be set for OFF (disabled) to prevent inadvertent activation  
during travel or storage. When set ON, the Pro Plus 2 will  
automatically Activate when immersed in water.  
• Press Both buttons simultaneously, release when SET: 2  
appears (4 seconds).  
• Press and release the Left (Advance) button, the Units  
screen appears with the set point flashing.  
• Press the Left (Advance) button 18 more times  
N2  
(14 more times after a dive in Digital Gauge Mode).  
• The graphics ACT and H20 appear with ON (or OFF)  
flashing (Fig. 42).  
• Press and release the Right (Select) button to toggle  
between ON and OFF.  
O2  
• Press the Left (Advance) button to accept the setting and  
revert to Surface Mode.  
ATR  
Fig.42-SetWaterActivation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
44  
WARNING: PriortodivingwiththeProPlus2,you  
mustalsoreadandunderstandtheOceanicDive  
ComputerSafetyandReferenceManualwhich  
providesImportantWarningsandSafetyRecom-  
mendationsaswellasgeneralproductinformation.  
PLAN and DIVE MODES  
Download from Www.Somanuals.com. All Manuals Search And Download.  
45  
PREDIVEPLANNINGSEQUENCETM  
Oceanic strongly recommends that you review the  
Pre Dive Planning Sequence (PDPS) prior to every  
dive to help you plan your dive as required to  
avoid exceeding no decompression or oxygen  
exposure limits.  
Depth  
NDL  
This is especially important for repetitive dives when the  
PDPS indicates adjusted dive times that are available for the  
next dive, based on residual nitrogen or oxygen accumula-  
tion (whichever is in control) following the last dive and  
surface interval.  
feet(meters) hours:mins  
30 (9)  
4:20 (4:43)  
2:17 (2:24)  
1:21 (1:25)  
40 (12)  
50 (15)  
60 (18)  
70 (21)  
80 (24)  
90 (27)  
100 (30)  
110 (33)  
120 (36)  
130 (39)  
140 (42)  
150 (45)  
160 (48)  
170 (51)  
180 (54)  
190 (57)  
:57  
:40  
:30  
:24  
:19  
:16  
:13  
:11  
:09  
:08  
:07  
:07  
:06  
:05  
(:59)  
(:41)  
(:32)  
(:25)  
(:20)  
(:17)  
(:14)  
(:11)  
(:09)  
(:08)  
(:07)  
(:06)  
(:06)  
(:05)  
The PDPS provides a sequence of theoretical dive times  
available for depths ranging from 30 feet (9 meters) to 190  
feet (57 meters) in 10 foot (3 meter) increments.  
During the sequence, no decompression times are only  
displayed for depths where there is at least 1 minute of  
theoretical dive time available at the depth, taking into  
account a descent and ascent rates of 60 feet (18 meters)  
per minute.  
NoDecompressionLimits  
foranAirDive  
(nodivemadeyet)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
46  
N2  
To access the Pre Dive Planning Sequence  
(while in the Surface mode) -  
FEET  
• Press the Left (Advance) button 1 time  
MAX  
• Press and release the Right (Select) button to advance  
through the Depths/Times, one screen at a time; and then  
to Surface Mode.  
• Press the Left (Advance) button to access Fly Mode.  
• The unit will revert to Surface Mode after 2 minutes if no  
button is pressed.  
O2  
ATR  
• With each depth displayed by the Pre Dive Planning  
Sequence, you will see either predicted no decompres-  
sion limits (NDLs) based upon your previous dive profiles  
(if calculated to be nitrogen controlled), or predicted  
oxygen tolerance limits (OTLs) based upon either a single  
dive exposure or your 24 hour accumulation of oxygen (if  
calculated to be oxygen controlled).  
Fig.43-NitrogenControl  
N2  
FEET  
• The Maximum Depth allowed for the PO2 Alarm and  
FO2 set points will also be displayed.  
MAX  
• Depths greater than the maximum depth that can be  
achieved with a PO2 of 1.60 ATA will not be displayed.  
• If the Tissue Loading Bar Graph is displayed (Fig. 43),  
that next dive is calculated to be controlled by nitrogen  
loading.  
• If the O2BG and O2 symbol are displayed (Fig. 44), it is  
calculated to be controlled by oxygen loading.  
O2  
O2  
ATR  
Fig.44-OxygenControl  
Download from Www.Somanuals.com. All Manuals Search And Download.  
47  
NOTE: TheProPlus2willstoreoxygenaccumula-  
tioncalculationsforupto10divesconductedduringa  
24hourperiod. Ifthemaximumlimitforoxygen  
loadinghasbeenexceededforthatday(24hour  
period),allofthesegmentsoftheO2bargraphwillbe  
displayedflashing.  
Depth/TimevalueswillnotappearuntiltheO2bar  
graphrecedesintothegreen(normal)zone(i.e.,your  
dailyoxygendosagedecreasesanamountequivalent  
totheamountaccumulatedduringthelatestdive  
completed).  
a
TISSUELOADINGBARGRAPH(Fig.45a)  
N2  
FEET  
As your Depth and Elapsed Dive Time increase, the Tissue  
Loading Bar Graph (TLBG) will add segments (green toward  
red) to represent the absorption of nitrogen.  
While ascending to shallower depths, the number of seg-  
ments displayed will begin to recede, offering a graphic  
representation of your multilevel diving capability.  
O2  
P SI  
0
5
1 0  
2 0  
ATR  
Fig.45-TissueLoadingBarGraph  
Download from Www.Somanuals.com. All Manuals Search And Download.  
48  
N2  
OXYGENACCUMULATIONBARGRAPH(Fig.46a)  
FEET  
If FO2 was set for a numerical value (nitrox), the Oxygen  
Accumulation (O2) Bar Graph will add segments (green  
toward red) to represent oxygen accumulation for that dive,  
or 24 hour period, whichever amount is greater.  
a
b
VARIABLEASCENTRATEINDICATOR(Fig.46b)  
O2  
P SI  
0
5
ATR  
1 0  
2 0  
The Variable Ascent Rate Indicator (VARI) shows how fast you  
are ascending. When you exceed the maximum recom-  
mended ascent rate for the depth you are at (see below), it  
will enter the red (Too Fast) zone and you will be alerted by  
the graphic TOO FAST, all segments of the bar graph  
flashing, an Audible alarm and the red LED Warning Indica-  
tor, unless set OFF (Fig. 47). The warnings will stop when  
your ascent rate is slowed.  
Fig.46-O2BarGraph&VARI  
N2  
FEET  
TOO  
FAS T  
• At depths greater than 60 feet (18 meters), maximum  
recommended rate is 60 feet/minute (18 meters/minute).  
• At depths of 60 feet (18 meters) and shallower, maxi-  
mum recommended rate is 30 feet/minute (9 meters/  
minute).  
O2  
P SI  
0
5
1 0  
2 0  
3 0  
ATR  
NOTE: Refertopage10foradditionalinformation  
pertainingtotheVariableAscentRateIndicator.  
Fig.47-AscentAlarm  
Download from Www.Somanuals.com. All Manuals Search And Download.  
49  
AIRTIMEREMAININGBARGRAPH  
N2  
FEET  
The Air Time Remaining Bar Graph (Fig. 48a) displays remain-  
ing breathing gas times of 60 minutes and less. As time  
remaining decreases, segments are removed from the bar  
graph from the right (green) toward the left (red) in increments  
of 5 minutes (from 60 to 0 minutes).  
O2  
P SI  
When the time decreases to the Alarm set point, the Audible  
Alarm will emit a beep and the red Warning LED will flash until  
acknowledged. You should begin a safe controlled ascent to  
the surface.  
0
5
ATR  
1 0  
2 0  
a
Fig.48-AirTimeRemaining  
BarGraph  
CONTROLOFDISPLAYS  
During No Decompression conditions, various displays of information (up to 4) are  
available. Each provides Depth, Dive Time Remaining, and additional information. The  
intent of this feature is to provide the diver control of how much information is on the  
display at any given time during the dive. The diver can change from one display to  
another as often as desired by pressing the Left (Advance) button, otherwise it does not  
change. An exception to this is during a 3 minute no decompression dive Safety Stop  
(described on page 53).  
Download from Www.Somanuals.com. All Manuals Search And Download.  
50  
During conditions in which Cautionary type information is displayed (e.g., No Decom-  
pression Safety Stop, Decompression, High PO2, etc.), there is a Main Display of impor-  
tant information relevant to the specific condition. The diver can access another display,  
but it will automatically revert to the Main Display after 3 seconds.  
To activate the Backlight during a dive, press the Right (Select) button.  
• The display will be illuminated as long as the button is depressed, plus it will remain  
illuminated for the Backlight Duration time that has been set (3 or 7 seconds).  
• The Backlight will not activate during a Low Battery condition or if the Backlight  
duration is set for 0.  
NODECOMPRESSIONDIVEMODE  
NoDeco  
Modeicon  
The Pro Plus 2 will enter the No Decompression Dive Mode  
when you descend deeper than 4 feet (1.2 meters).  
N2  
FEET  
NoDecompressionDiveMode-Display#1(Fig.49)  
Information includes Current Depth, Dive Time Remaining and  
icon, Tank Pressure, and the applicable bar graphs.  
• Press and release the Left (Advance) button to view Display  
#2.  
O2  
P SI  
0
5
1 0  
2 0  
ATR  
Fig. 49 -NoDeco#1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
51  
N2  
NoDecompressionDiveMode-Display#2(Fig.50)  
FEET  
Information includes Current Depth, Max Depth for that dive  
and icon, Elapsed Dive Time and icon, Dive Time Remaining  
and icon, Tank Pressure, and the applicable bar graphs.  
MAX  
• Press and release the Left (Advance) button again to view  
Display #3.  
O2  
P SI  
NoDecompressionDiveMode-Display#3(Fig.51)  
0
5
ATR  
1 0  
2 0  
Fig. 50 -NoDeco#2  
Information includes Current Depth, Temperature, Time of  
Day, Dive Time Remaining and icon, Tank Pressure, and the  
applicable bar graphs.  
N2  
FEET  
• Press and release the Left (Advance) button again to view  
Display #4 (only if FO2 is set for a numerical value -  
Nitrox), or revert to Display #1 (if FO2 is set for Air).  
(lowerportion sameasFig.50)  
Fig.51-NoDeco#3  
NoDecompressionDiveMode-Display#4(Fig.52)  
N2  
FEET  
Information includes - Current Depth, current value of PO2  
(if a nitrox dive) and PO2 graphic, Dive Time Remaining and  
icon, Tank Pressure, and applicable bar graphs.  
(lowerportion sameasFig.50)  
Fig. 52 -NoDeco#4  
• press the Left (Advance) button to view Display #1.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
52  
N2  
NoDecompressionDiveMode-SecondaryDisplay(Fig.53)  
FEET  
To view, press the Left (Advance) button for 2 seconds.  
Information includes Current Depth, Air Time Remaining  
(and Mode icon), and applicable bar graphs.  
The Display will remain on the screen for 3 seconds then  
revert back to the Display previously selected (#1, 2, 3, 4).  
O2  
P SI  
0
5
1 0  
2 0  
ATR  
NoDecompressionDiveMode-SAFETYSTOP(Fig.54)  
Fig.53-NoDecoSecondary  
Upon ascending to 20 feet (6 meters) on any No Decompres-  
sion dive in which depth exceeded 30 feet (9 meters), a short  
beep will be emitted and a Safety Stop at 15 feet (4.5  
meters) will appear on the display with a 3 minute count-  
down timer that counts down from 3:00 to :00 (min:sec).  
N2  
FEET  
The Safety Stop will be displayed until the countdown times  
out, or the diver descends below 30 feet (10 meters), or the  
diver surfaces. There is no Penalty if the diver surfaces prior  
to completing the Safety Stop.  
O2  
P SI  
Information includes Current Depth, Stop Depth (15 feet or  
4.5 meters), Stop Bar icon, Countdown Timer, Dive Time  
Remaining, Cylinder Pressure, and applicable bar graphs.  
0
5
ATR  
1 0  
2 0  
3 0  
Fig.54-NoDecoSafety Stop  
Download from Www.Somanuals.com. All Manuals Search And Download.  
53  
DECOMPRESSIONDIVEMODE  
The Pro Plus 2 is designed to help you by providing a repre-  
sentation of how close you are to entering Decompression.  
Decompression Dive Mode (Fig. 55) activates when theoreti-  
cal No Decompression time/depth limits are exceeded.  
Upon entering Decompression Mode, the Audible Alarm will  
sound and the red LED Warning Light will flash for 10  
seconds (unless set OFF), or until acknowledged.  
DecoModeicon  
• Press and hold the Left (Advance) button for 2 seconds to  
acknowledge/silence the Audible Alarm (unless set OFF).  
• The UP Arrow and Deco Bar will flash if you are greater  
than 10 feet (3 m) deeper than the Required Stop Depth.  
• Once you are within 10 feet (3 m) of, and below, the  
Stop Depth, both Arrows and the Bar appear solid.  
DECO  
N2  
FEET  
TotalAscentTime  
a
Total Ascent Time (Fig. 55a) includes Stop Times required at  
all required decompression ceilings and vertical Ascent Time  
calculated at 60 feet (18 meters) per minute for depths  
deeper than 60 feet (18 meters), and 30 feet (9 meters) per  
minute for depths of 60 feet (18 meters) and shallower.  
O2  
P SI  
0
5
ATR  
1 0  
Fig.55-EntryintoDecoMode  
Download from Www.Somanuals.com. All Manuals Search And Download.  
54  
ManagingDecompressionStops  
To fulfill your decompression obligation, you should make a  
safe controlled ascent to a depth slightly deeper than, or  
equal to, the Required Ceiling Stop Depth indicated (Fig.  
56a) and decompress for the Stop Time indicated (Fig. 56b).  
The amount of decompression credit time that you receive is  
dependent on depth, with slightly less credit given the deeper  
you are. You should stay slightly deeper (Fig. 56c) than the  
Required Stop Depth indicated until the next shallower Stop  
Depth appears. Then, you can slowly ascend to, but not  
shallower than that indicated ceiling Stop Depth.  
c
• While in Decompression Mode, the Pro Plus 2 will  
automatically revert to the Main Display after 3 seconds  
unless the Left (Advance) button is pressed to view an-  
other display of information.  
DECO  
N2  
FEET  
b
DecompressionDiveMode-MainDisplay#1(Fig.56)  
Information includes - Current Depth, required Ceiling Stop  
Depth and Time, Total Ascent Time (and Mode icon), both  
Arrows and the Deco Bar, and the applicable bar graphs.  
a
O2  
P SI  
0
5
1 0  
2 0  
ATR  
• Press and release the Left (Advance) button to view  
Display #2.  
Fig. 56-Deco#1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
55  
DECO  
N2  
DecompressionDiveMode-Display#2(Fig.57)  
Information includes - Current Depth, Maximum Depth for  
that dive (and icon), Elapsed Dive Time (and icon), Total  
Ascent Time (and Mode icon), both Arrows and the Deco  
Bar, and the applicable bar graphs.  
FEET  
MAX  
• Press and release the Left (Advance) button again to view  
Display #3.  
O2  
P SI  
DecompressionDiveMode-Display#3(Fig.58)  
Information includes - Current Depth, Temperature, Time of  
Day (and icon), Total Ascent Time, both Arrows and the  
Deco Bar, and the applicable bar graphs.  
0
5
ATR  
1 0  
2 0  
Fig. 57-Deco#2  
DECO  
N2  
FEET  
• Press and release the Left (Advance) button again to view  
- Display #4 (only if FO2 is set for a numerical value -  
Nitrox), or revert to Display #1 (if FO2 is set for Air).  
(lowerportionsameasFig.57)  
Fig. 58-Deco#3  
DecompressionDiveMode-Display#4(Fig.59)  
Information includes - Current Depth, the current value of  
PO2 (if a nitrox dive) and PO2 graphic, Total Ascent Time,  
Mode icon, both Arrows and the Deco Bar, and applicable  
bar graphs.  
DECO  
N2  
FEET  
• Press the Left (Advance) button to view Display #1.  
(lowerportionsameasFig.57)  
Fig. 59-Deco#4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
56  
VIOLATIONMODES  
While in Violation Modes the Alternate Displays previously described can be accessed  
using the Left (Advance) button, and the Backlight can be activated using the Right  
(Select) button.  
• The Pro Plus 2 will automatically revert to the Main Display after 3 seconds unless  
the Left (Advance) button is pressed to view another display of information.  
ConditionalViolationMode(Fig.60)  
If you ascend shallower (Fig. 60a) than a Required Decompression Ceiling  
Stop Depth displayed (Fig. 60b), the Down Arrow, Deco Bar, and Total Ascent Time  
will flash until you descend below the Required Stop Depth. Also displayed will be  
a
Current Depth and applicable bar graphs.  
DECO  
N2  
FEET  
Unless set OFF, the Audible Alarm will emit a continuous tone  
and the red LED Warning Light will flash for 10 seconds, or  
until acknowledged by pressing the Left (Advance) button.  
b
If you descend below the required decompression ceiling  
before 5 minutes have elapsed, the Pro Plus 2 will continue to  
function in Decompression Dive Mode. In this case, no off-  
gassing credit will be given, and for each minute above the  
ceiling 11/2 minutes of Penalty Time will be added to  
Required Stop Time.  
O2  
P SI  
0
5
ATR  
1 0  
2 0  
Fig.60 - ConditionalViolation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
57  
The added Penalty (decompression) Time will have to be  
'worked off' first, before obtaining off-gassing credit. Once  
the Penalty Time is worked-off, and off-gassing credit begins,  
required decompression Stop Depths and Time will decrease  
toward zero. The Tissue Loading Bar Graph will recede into  
the yellow Caution Zone and the Pro Plus 2 will revert to the  
No Decompression Dive Mode.  
NOTE: Uponentryintothefollowing3Delayed  
ViolationModes,theredLEDwarninglightwilllight  
andtheAudibleAlarmwillsound,evenif SetOFF.  
TheAlarmcannotbeacknowledged(silenced)by  
pressingtheLeft(Advance)button.  
DelayedViolationMode#1(Fig.61)  
(This is a continuation of a Conditional Violation.)  
DECO  
N2  
FEET  
If you remain above the Required Ceiling Stop  
Depth for more than 5 minutes, the full Tissue Loading  
Bar Graph and Total Ascent Time will flash until you descend  
below the Required Stop Depth.  
O2  
P SI  
0
5
1 0  
2 0  
ATR  
Fig.61- DelayedViolation#1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
58  
DECO  
N2  
DelayedViolationMode#2(Fig.62)  
FEET  
The Pro Plus 2 cannot calculate decompression times for Stop  
Depths much greater than 60 feet (18 meters) and offers no  
indication of how much time spent underwater would result in  
the need for a greater Stop Depth.  
If the Decompression obligation requires a Ceiling  
Stop Depth between 60 feet (18 meters) and 70  
feet (21 meters), the Tissue Loading Bar Graph will flash.  
Total Ascent Time will still be displayed.  
O2  
P SI  
0
5
ATR  
1 0  
2 0  
3 0 4 0  
5 0  
Fig.62- DelayedViolation#2  
You must ascend to just deeper than, and stay as close as  
possible to, 60 feet (18 meters) without causing the Total  
Ascent Time display to flash. When the Required Stop Depth  
indicates 50 FT/ 15 M, etc., you can ascend to those depths  
and continue decompressing.  
N2  
FEET  
DelayedViolationMode#3(Fig.63)  
MAX  
If you descend deeper than 330 feet (99.9  
meters), the accumulated Tissue Loading Bar Graph seg-  
ments will flash, and the Current Depth and Max Depth  
displays will only indicate 3 dashes ( - - -).  
O2  
P SI  
0
5
ATR  
1 0  
2 0  
Fig.63 -DelayedViolation#3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
59  
Upon ascending above 330 feet (99.9 meters), the Current  
Depth display will be restored, however Max Depth will only  
display 3 dashes for the remainder of that dive. The Log for  
that dive will display 3 dashes as the Max Depth achieved.  
ImmediateViolationModeandGaugeMode  
During a Dive, if a ceiling much greater than 60 feet (18  
meters) is required, an Immediate Violation Mode will  
be entered. This situation would be preceded by entering  
Delayed Violation Mode #2.  
The Pro Plus 2 would then operate with limited functions in  
Gauge Mode during the remainder of that dive and for 24  
hours after surfacing. Gauge Mode turns the Pro Plus 2 into  
a digital instrument without any decompression or oxygen  
monitoring functions.  
DECO  
N2  
FEET  
MAX  
Only Current Depth, Max Depth, Elapsed Dive Time, and the  
bar graphs will be displayed (Fig. 64). The full Tissue Load-  
ing Bar Graph and O2 Bar Graph will both flash as a warn-  
ing of this condition.  
O2  
O2  
P SI  
0
5
1 0  
ATR  
• Press the Left (Advance) button to view Temperature and  
Time of Day.  
• Press the Right (Select) button to activate the Backlight.  
Fig.64 -ImmediateViolation  
GaugeMode(underwater)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
60  
The Pro Plus 2 will also enter an Immediate Violation  
Mode 5 minutes after reaching the surface from a dive in  
which a Delayed Violation occurred.  
On the surface, Gauge Mode displays the Dive Number,  
Temperature, Time of Day, and Surface Interval (Fig. 65). It  
does not provide the FO2, PDPS, or Time to Fly and  
Desaturate features.  
The countdown timer that appears when you try to access  
Time to Fly does not represent 'Time to Fly'. It is only pro-  
vided to inform you of the time remaining before normal Pro  
Plus 2 operation can resume with full features and functions.  
DECO  
This condition is a Permanent Violation, and in the  
event that a dive is made during the 24 hour  
period, a full 24 hour surface interval must then  
be served before all functions are restored.  
N2  
O2  
Responnsible  
O2  
P SI  
ATR  
Fig.65 -ImmediateViolation  
GaugeMode(aftersurfacing)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
61  
N2  
HIGHPO2DIVEMODE  
FEET  
When partial pressure of oxygen (PO2) becomes equal to, or  
greater than, 1.40 ATA, or 0.2 ATA less than the  
PO2 Alarm set point (a user setting); the red LED  
warning will flash, the Audible Alarm will sound (unless set  
OFF), and the current PO2 value, PO2 symbol, O2 segment  
of the O2 Bar Graph, and UP Arrow will appear on the  
Main Display as a warning until PO2 decreases. Current  
Depth and Dive Time Remaining will also be displayed (Fig.  
66).  
O2  
O2  
0
P SI  
5
1 0  
2 0  
3 0  
ATR  
Fig.66-PO2Warning  
If PO2 continues to increase, the value displayed will in-  
crease toward a maximum value of 5.00 ATA in increments  
of .01 ATA. When it reaches a value of 1.60 ATA, or the  
PO2 Alarm set point (a user setting), the Audible  
Alarm will sound (unless set OFF) and the current PO2 value,  
PO2 symbol, O2 segment of the O2 Bar Graph, and UP  
Arrow will flash as a warning until PO2 decreases (Fig. 67).  
N2  
FEET  
• Press the Left (Advance) button to view the Alternate  
Displays.  
• The unit will revert to the Main Display after 3 seconds.  
• Press the Right (Select) button to activate the Backlight.  
O2  
O2  
P SI  
0
5
1 0  
2 0  
3 0  
ATR  
Fig.67-PO2Alarm  
Download from Www.Somanuals.com. All Manuals Search And Download.  
62  
HIGHOXYGENACCUMULATION  
The O2 Bar Graph displays either oxygen accumulated  
during that nitrox dive, or during the repetitive nitrox dives  
you conduct during that 24 hour period, whichever of the  
two is greater at that time.  
The yellow Caution Zone of the O2 Bar Graph offers you a  
convenient way to consistently monitor how close you are  
coming to the limits of oxygen exposure. Use it as a  
visual reference to place a wider margin of protec-  
tion between you and the Limits.  
If the theoretical amount of oxygen accumulated equals, or  
exceeds, the limit for a single exposure, or the exposure limit  
for a 24 hour period, Oxygen Dive Time Remaining becomes  
zero (0:00) and the O2 Bar Graph will enter the red O2  
(Danger) Zone (Fig. 68). The red LED warning will flash,  
the Audible Alarm will sound (unless set OFF), and the UP  
Arrow and the full O2 Bar Graph will flash as a warning until  
the level of oxygen decreases below the limit.  
N2  
FEET  
O2  
• Press the Left (Advance) button to view the Alternate  
Displays.  
• The unit will revert to the Main Display after 3 seconds.  
• Press the Right (Select) button to activate the Backlight.  
O2  
O2  
0
P SI  
5
1 0  
ATR  
Fig.68-HighO2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
63  
N2  
USERSETDIGITALGAUGEMODE  
FEET  
When Digital Gauge Mode is set for ON, the Pro Plus 2 will  
operate as a Digital Depth Gauge/Timer without performing  
nitrogen and oxygen calculations (Fig. 69).  
MAX  
Time of Day and Elapsed Dive Time will also be displayed.  
O2  
• Press the Left (Advance) button to view Air Time Remain-  
ing for 3 seconds in place of Elapsed Dive Time (Fig. 70).  
P SI  
0
5
ATR  
1 0  
Fig.69 -UserSetGaugeMode  
UNEXPECTEDLOSSOFDISPLAYEDINFORMATION  
If your Pro Plus 2 stops working for any reason, it is impor-  
tant that you have anticipated this possibility and are pre-  
pared for it. This is an important reason for not pushing the  
no decompression and oxygen accumulation limits, and a  
critical reason to avoid entering decompression. If you dive  
in situations where your trip would be ruined or your safety  
would be jeopardized by losing the use of your Pro Plus 2, a  
backup instrument system is highly recommended.  
N2  
FEET  
MAX  
O2  
P SI  
0
5
1 0  
ATR  
Fig.70 -UserSetGaugeMode  
(AirTimeRemaining)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
64  
WARNING: PriortodivingwiththeProPlus2,you  
mustalsoreadandunderstandtheOceanicDive  
ComputerSafetyandReferenceManualwhich  
providesImportantWarningsandSafetyRecom-  
mendationsaswellasgeneralproductinformation.  
POST DIVE MODES  
Download from Www.Somanuals.com. All Manuals Search And Download.  
65  
POSTDIVESURFACEMODE  
When you ascend to 3 feet (1 meter) or shallower, the Pro  
Plus 2 will enter Surface Mode and begin counting your  
surface interval.  
TRANSITIONPERIOD  
The first 10 minutes is, in affect, a Transition Period during  
which time the following information is displayed (Fig. 71):  
• Number of that dive (during that day)  
• Battery Consumption Indicator  
• Temperature (ambient)  
• Time of Day and icon  
N2  
• Surface Interval time (colon flashing) and icon (flashing)  
• Tissue Loading Bar Graph indicating current nitrogen  
loading  
• O2 Bar Graph indicating current oxygen accumulation (if  
nitrox)  
• During the Transition Period, Log Mode can be accessed.  
No other modes (e.g., PDPS, Fly, Desat, Set, PC) are  
accessible.  
O2  
P SI  
ATR  
• Press the Right (Select) button to activate the Backlight.  
Fig.71-TransitionPeriod  
Download from Www.Somanuals.com. All Manuals Search And Download.  
66  
To view that dive's Log (Fig. 72) -  
Refer to page 64 for a description of the Log Mode displays.  
• Press the Left (Advance) button 1 time.  
• Press the Right (Select) button 1 time to view the Nitro-  
gen data screen.  
• Press the Right (Select) button again to view the Oxygen  
data screen (if a nitrox dive).  
• Press Both buttons simultaneously for 2 seconds to return  
to Surface Mode.  
• The unit will revert to Surface Mode after 2 minutes if no  
button is pressed.  
Log Data will not be stored in the unit's memory until the 10  
minute Transition Period on the surface is completed.  
N2  
Once 10 minutes have elapsed, the Surface Mode icon and  
Surface Interval time display colon stop flashing indicating  
that the Dive and Transition Period are completed, and a  
subsequent descent will be considered a new dive.  
If you descend during the 10 minute Transition Period, time  
underwater will be considered a continuation of that dive.  
The time at the surface (if less than 10 minutes) will not be  
added as Dive Time.  
O2  
ATR  
Fig.72-LogMode  
(duringTransitionPeriod)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
67  
N2  
AFTER THE TRANSITION PERIOD (THE FIRST 2 HOURS)  
For the remainder of the first 2 hours after surfacing,  
information will continue to be displayed as Surface Mode  
(Fig. 73) and you will have full access to other modes (e.g.,  
PDPS, Fly, Desat, Log, Set, etc.).  
ToactivatetheBacklight-  
O2  
• Press the Right (Select) button.  
ATR  
ToaccessthePreDivePlanningSequence-  
(Also refer to page 46)  
Fig.73-SurfaceMode(>10min)  
• Press the Left (Advance) button 1 time (while in Surface  
Mode).  
• Press and release the Right (Select) button to advance  
through the sequence of available 'adjusted' depths and  
times one screen at a time.  
N2  
FEET  
• The unit will revert to Surface Mode after 2 minutes,  
unless the Left (Advance) button is pressed to access the  
Fly Mode.  
The Pre Dive Planning Sequence now shows adjusted No  
Decompression Limits (Fig. 74) based on residual nitrogen  
and accumulated oxygen calculated to be remaining from  
the previous dives.  
O2  
ATR  
Fig.74- AdjustedNDLs  
Download from Www.Somanuals.com. All Manuals Search And Download.  
68  
N2  
ToaccesstheTimetoFlyCountdown(whileinSurfaceMode)  
• Press the Left (Advance) button 2 times.  
• The Time to Fly counter (Fig. 75) begins counting down 10  
minutes after surfacing from a dive (after the Transition  
Period) displaying the word 'FLY' and a countdown that  
begins at 23:50 (hr:min) and counts down to 0:00  
(hr:min).  
• If a Violation occurred during the dive a single dash ( - )  
will appear instead of the letters FLY.  
O2  
• The unit will revert to Surface Mode after 2 minutes, unless  
the Left (Advance) button is pressed to access the Desat  
Time Countdown.  
ATR  
Fig.75-TimetoFly  
ToaccesstheTimetoDesaturationCountdown  
(whileinSurfaceMode)  
N2  
• Press the Left (Advance) button 3 times.  
• The unit will revert to Surface Mode after 2 minutes, unless  
the Left (Advance) button is pressed to access the Log  
Mode.  
• If a Violation occurred during the dive, Desaturation Time  
will not be displayed.  
The Time to Desaturate counter (Fig. 76) provides calculated  
time for tissue desaturation at sea level. The countdown starts  
10 minutes after surfacing at 23:50 (hr:min) maximum and  
counts down to 0:00 (hr:min).  
O2  
ATR  
Fig.76-TimetoDesaturate  
Download from Www.Somanuals.com. All Manuals Search And Download.  
69  
LOGMODE  
Information from your latest 24 dives is stored in the Log for viewing. The first dive of a  
new 'activation period' will be #1, then #2, etc. After 24 dives are accumulated, each  
subsequent dive will overwrite the oldest dive in the Log (i.e., the most recent dive deletes  
the oldest). Log information will not be lost when the battery is removed, but factory  
service will delete data.  
Dives are displayed in a reverse sequence that starts with the dive most recently recorded  
back to the oldest of the 24 dives stored. Thus, your most recent dive will always be the  
first shown in the sequence. Each dive has three log screens - Date/Time started, Nitro-  
gen data, and Oxygen data (not displayed if the dive included a Violation). Dives are  
identified by the Date/Time started and 'number' that day.  
Button Control in Log Mode -  
• The Left (Advance) button is used to access a specific dive's Log.  
• HINT: To bypass a dive's Log and search for another in the sequence, press the Left  
(Advance) button repeatedly. Do not press the Right (Select) button until you find the  
dive Log you wish to view.  
• The Right (Select) button is used to view the second and third screens (Nitrogen and  
Oxygen related data) for that dive.  
• To return to Surface Mode at any time while in Log Mode, press Both buttons simulta-  
neously for 2 seconds.  
• The unit will automatically revert to Surface Mode after 2 minutes if no button is  
pressed while in the Log Mode.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
70  
N2  
To access the Log Mode and view the First Screen  
(While in Surface Mode) -  
• Press the Left (Advance) button 4 times.  
• The First Screen of the most recent dive will appear  
displaying (Fig. 77) -  
• The Log Mode icon  
• Dive Number  
• Time/ Date the dive started  
O2  
ATR  
To view the Second Screen of the Log (Nitrogen  
data) -  
Fig.77-Log(Date/Time)  
• Press the Right (Select) button 1 time (while viewing the  
Time/Date). Displayed will be (Fig. 78) -  
• Log Mode icon  
• Maximum Depth - reached during the dive (and icon)  
• Temperature - minimum during the dive (and icon)  
• Surface Interval - prior to that dive (and icon)  
• Elapsed Dive Time (and icon)  
N2  
FEET  
MAX  
• Variable Ascent Rate Indicator - showing the maximum  
ascent rate maintained for 4 consecutive seconds during  
the dive  
• Tissue Loading Bar Graph - showing tissue nitrogen  
loading at the time you surfaced at the end of the dive.  
Also, the segment that reflects the maximum loading will  
appear flashing.  
O2  
ATR  
Fig.78-Log(NitrogenData)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
71  
N2  
To view the Third Screen of the Log (Oxygen data)  
• Press the Right (Select) button 1 time (while viewing the  
Nitrogen data). Displayed will be (Fig. 79)-  
• Log Mode icon  
• FO2 graphic and value set for that dive  
• Maximum PO2 level reached during that dive and PO2  
graphic  
• O2 Bar Graph - showing oxygen loading at the end of  
the dive  
O2  
ATR  
Fig.79-Log (OxygenData)  
To access the first screen of the previous dive's Log  
• Press the Left (Advance) button 1 time.  
AFTER THE FIRST 2 HOURS  
N2  
Two hours after the last dive, Surface Mode will no longer be  
displayed, the Time to Fly and Desaturation countdowns will  
be displayed alternately for 3 seconds each until they count  
down to 0:00 or another dive is made.  
To access other modes or enter settings -  
• Press Either button to return to Surface Mode (Fig. 80).  
• The unit will again revert to the Time to Fly and  
Desaturation countdowns after 2 hours, if no button is  
pressed.  
O2  
ATR  
Fig.80-SurfaceMode  
Download from Www.Somanuals.com. All Manuals Search And Download.  
72  
N2  
WET CONTACTS  
If the graphic H2O appears during the Fly Mode (Fig. 81)  
and Desaturation Mode (Fig. 82) countdowns that alternate  
2 hours after a dive, it is an indication that the water activa-  
tion contacts are bridged (still wet) and the unit must be  
rinsed in fresh water and thoroughly dried.  
O2  
One contact is located between the control buttons and the  
other at the base of the Right (Select) button's stem.  
ATR  
Fig.81-FlyMode  
(ActivationContactsWet)  
• Once the unit is dry, the graphic H2O will disappear  
from the display.  
• If the unit is not cleaned and dried prior to the count-  
downs reaching 0:00 (hr:min), or making another dive, it  
will shut off then automatically reactivate.  
N2  
• The graphic H2O would then appear in place of Dive  
Number when Surface Mode is displayed during the  
Surface Mode.  
O2  
• If no dive is made, the unit would shut off after 2 hours,  
then automatically reactivate again, repeating the action  
until cleaned and dried.  
ATR  
Fig.82-DesaturationMode  
(activationcontactswet)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
73  
DOWNLOADINGDATATOAPC  
Using special linking hardware, dive data can be down-  
loaded (copied) from your Pro Plus 2 to an IBM compatible  
PC program running on a Windows® operating system.  
Compatibility requirements and instructions are provided  
with the optional download package that is available from  
your Authorized Oceanic Dealer. The software program  
provides dive tabular and graphic profile data sampled  
throughout the dives at the interval you set.  
NOTE: Ensurethatthedownloadproductthatyou  
acquireiscompatiblewiththeProPlus2andthePC  
equipmentthatyouwillbeusing.  
N2  
The Interface Cable will be connected to the Data Port  
located on the side of the Pro Plus 2 housing.  
Prior to attempting to download data from your  
Pro Plus 2, refer to the instructions provided in the  
User Manual that is incorporated into the CD for  
the download package.  
O2  
Refer to page 28 of this manual for instructions  
regarding access to PC Interface (Fig. 83).  
ATR  
Fig.83-PCInterface  
Download from Www.Somanuals.com. All Manuals Search And Download.  
74  
WARNING: PriortodivingwiththeProPlus2,you  
mustalsoreadandunderstandtheOceanicDive  
ComputerSafetyandReferenceManualwhich  
providesImportantWarningsandSafetyRecom-  
mendationsaswellasgeneralproductinformation.  
SIMULATOR (DEMO) MODE  
Download from Www.Somanuals.com. All Manuals Search And Download.  
75  
N2  
SIMULATORMODE  
This mode provides you with the ability to practice various  
dive mode scenarios and computer functions while observing  
the various displays.  
DEMO  
• The SmartGlo Backlight functions as it normally does.  
• At any time while in Simulator Mode, pressing and  
holding both buttons simultaneously for 2 seconds shall  
revert operation to real Surface Mode.  
O2  
• The real set points entered into the Pro Plus 2 do not  
affect the operation of the Simulator which has its own  
settings that allow Digital Gauge Mode to be set ON or  
OFF, calculations to be cleared, and FO2 to be set.  
ATR  
Fig.84-SimulatorMode  
AccessandSetup(whileinrealSurfaceMode)  
N2  
• Press and hold Both buttons for 6 seconds.  
• Release the buttons during the 2 second window when  
SIM and DEMO appear (Fig. 84).  
DEMO  
• Press and release the Left (Advance) button to access SIM  
MODE. The graphics GAU and DEMO will appear  
with OFF (or ON) flashing (Fig. 85).  
• When set ON, the SIMULATOR will operate as the unit  
would in User Set Digital Gauge Mode only displaying  
Depth and Elapsed Dive Time. When set OFF, it oper-  
ates as it would as an Air or Nitrox computer.  
O2  
ATR  
Fig.85 -SetDemoGauge  
Download from Www.Somanuals.com. All Manuals Search And Download.  
76  
N2  
• Press and release the Right (Select) button to toggle  
between Gauge ON and OFF.  
DEMO  
• Press and release the Left (Advance) button to accept the  
setting and advance to DEMO: NI-O2 with CUR (or  
NEW) flashing (Fig. 86).  
• When set for NEW, calculations are based on zero  
residual nitrogen and oxygen loading (a clean dive).  
• When set for CUR, calculations take into consideration  
any residual nitrogen and oxygen remaining from previ-  
ous 'actual' dives.  
• Subsequent (repetitive) Simulated dives do not take into  
consideration previous Simulated dives.  
O2  
ATR  
Fig.86-SetDemoCalibration  
• Press and release the Right (Select) button to toggle  
between NEW and CUR.  
N2  
• Press and release the Left (Advance) button to accept the  
setting and advance to DEMO: PSI (or bAr) with OFF,  
CYL, or 3000 flashing (Fig. 87).  
DEMO  
• When set for OFF, no calculations are made and no  
displays appear relating to tank pressure.  
• When set for CYL, Air Time Remaining is calculated  
based upon actual pressure of the tank that the Pro Plus 2  
is connected to.  
• When set for 3000, Air Time Remaining is calculated  
based upon a simulated tank pressure of 3000 psi.  
O2  
P SI  
ATR  
Fig.87 -SetPressureOn/Off  
Download from Www.Somanuals.com. All Manuals Search And Download.  
77  
N2  
• Press and release the Right (Select) button to toggle  
between Pressure ON and OFF.  
• Press and release the Left (Advance) button to accept the  
setting and revert to Demo Surface Mode with the  
DEMO icon flashing.  
DEMO  
• Press and release the Left (Advance) button to access  
DEMO: FO2 with the DEMO icon and Air (or a numeri-  
cal value) flashing (Fig. 88).  
O2  
• Press and hold the Right (Select) button to scroll from Air  
to 21% through 32% in 1% increments. The scroll will  
stop when the button is released, or at 32%.  
ATR  
Fig.88-SetDemoFO2  
• Press and hold the Right (Select) button again to scroll  
from 32% through 50% in 1% increments, then to Air.  
The scroll will stop when the button is released, or at Air.  
• Note that pressing and releasing the Right (Select) button  
advances the FO2 setting from AIR to 21 through 50, in  
increments of 1% per button depression (no scroll).  
• Press and release the Left (Advance) button to accept the  
setting and revert to Demo Surface Mode with the  
DEMO icon flashing.  
N2  
DEMO  
DemoDiveMode  
O2  
P SI  
• Press and hold the Right (Select) button for 2 seconds to  
access Demo Dive Mode. The DOWN Arrow will  
appear flashing (Fig. 89).  
0
5
ATR  
1 0  
2 0  
3 0 4 0  
5 0 6 0  
Fig.89-DemoDiveMode  
Download from Www.Somanuals.com. All Manuals Search And Download.  
78  
N2  
Descending  
Hint: Quick taps (<2 seconds) of the Right (Select) button  
starts/stops Descents and Time Acceleration. Quick taps of  
the Left (Advance) button accesses Alternate displays.  
DEMO  
• While the Down Arrow is flashing, tap (press/release)  
the Right (Select) button to begin a Descent at a rate of  
5 feet (1.5 meters) per real second.  
• Tap (press/release) the Right (Select) button during the  
Descent to stop the Descent.  
O2  
P SI  
0
5
ATR  
1 0  
2 0  
3 0 4 0  
5 0 6 0  
• Tap (press/release) the Left (Advance) button during the  
Descent, or stopped, to access the Alternate Displays.  
• Press and hold the Right (Select) button for 4 seconds to  
access Time Acceleration. The small clock icon begins  
flashing (Fig. 91).  
• Tap (press/release) the Right (Select) button to increase  
Elapsed Dive Time 1 minute per real time second.  
• Tap (press/release) the Right (Select) button during Time  
Acceleration to restore normal time rate of 1 second per  
real second.  
Fig.90 -DiveModeSetup  
N2  
FEET  
MAX  
DEMO  
• To start an Ascent, tap (press/release) the Right (Select)  
button to stop a Descent.  
O2  
P SI  
• To revert operation to real Surface Mode, press and hold  
Both buttons simultaneously for 2 seconds.  
0
5
ATR  
1 0  
2 0  
Fig.91 -TimeAcceleration  
Download from Www.Somanuals.com. All Manuals Search And Download.  
79  
N2  
Ascending  
FEET  
Rate of Ascent can be set for 1 or 3 feet per seconds (fps)  
prior to start of the Ascent. To change the Rate during an  
Ascent, it must first be stopped.  
DEMO  
• Pressing and holding the Left (Advance) button for 2  
seconds will access an Ascent Rate of 1 fps. The green  
and yellow segments of the VARI and small UP Arrow  
will appear with the Arrow flashing (Fig. 92).  
O2  
P SI  
0
5
1 0  
• To set the Rate at 3 fps (Too Fast), press and hold the Left  
(Advance) button for 2 seconds. The red segment of the  
VARI will appear together with the graphic TOO FAST  
and an outer Arrow around the small one (Fig. 93). The  
Alarm will sound and red LED will flash.  
ATR  
Fig.92-Ascentsetat1fps  
• Pressing and holding the Left (Advance) button for 2  
seconds will return the Rate to 1 fps.  
• Tap (press/release) the Left (Advance) button while the  
Arrow is flashing to begin an Ascent at the Rate set.  
• Tap (press/release) the Left (Advance) button during the  
Ascent to stop the Ascent.  
N2  
FEET  
TOO  
FAS T  
DEMO  
• Tap (press/release) the Left (Advance) button while the  
Ascent is stopped to access Alternate displays.  
• Press and hold the button for 2 seconds while the Ascent  
is stopped to restart the Ascent.  
• Pressing and holding both Buttons simultaneously for 2  
seconds will revert operation to real Surface Mode.  
O2  
P SI  
0
5
ATR  
1 0  
2 0  
Fig.93 -Ascentsetat3fps  
Download from Www.Somanuals.com. All Manuals Search And Download.  
80  
SurfacingfromaSimulatedDive  
• The Simulator will enter Demo Surface Mode (Fig.  
94) upon ascending to 3 feet (1 meter) or shallower.  
• Press and hold the Right (Select) button for 4 seconds to  
access Time Acceleration. The small Time Clock icon will  
begin flashing.  
• Tap (press/release) the Right (Select) button to increase  
Surface Interval Time 1 minute per real time second.  
• Tap (press/release) the Right (Select) button during the  
Time Acceleration to restore the time rate to normal.  
• FO2 can be set 10 minutes after surfacing.  
• Tap (press/release) the Left (Advance) button to access  
FO2 Set Mode. The FO2 graphic and value previously  
set will appear with the value flashing.  
N2  
• Press/release the Right (Select) button to increase the  
FO2 value in increments of 1%, or press and hold it to  
scroll.  
DEMO  
• Tap (press/release) the Left (Advance) button to accept  
the FO2 setting and return to Demo Surface Mode.  
• To make another dive, press and hold the Right (Select)  
button for 2 seconds to access Dive Mode.  
• To revert operation to real Surface Mode, press and hold  
Both buttons simultaneously for 2 seconds.  
O2  
P SI  
ATR  
Fig.94 -DemoSurfaceMode  
Download from Www.Somanuals.com. All Manuals Search And Download.  
81  
Be a -  
RESPONSIBLE DIVER  
at all times.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
82  
WARNING: PriortodivingwiththeProPlus2,you  
mustalsoreadandunderstandtheOceanicDive  
ComputerSafetyandReferenceManualwhich  
providesImportantWarningsandSafetyRecom-  
mendationsaswellasgeneralproductinformation.  
GENERAL  
Download from Www.Somanuals.com. All Manuals Search And Download.  
83  
CAREANDCLEANING  
Protect your Pro Plus 2 from shock, excessive temperatures,  
chemical attack, and tampering. Protect the lens against  
scratches with a transparent Instrument Lens Protector. Small  
scratches will naturally disappear underwater.  
• Soak and rinse the Pro Plus 2 in fresh water at the end of  
each day of diving, and check to ensure that the areas  
around the Low Pressure (Depth) Sensor (Fig. 95a),  
Download Interface Port (Fig. 95b), and Buttons are free  
of debris or obstructions.  
• To dissolve salt crystals, use lukewarm water or a slightly  
acidic white vinegar/water bath. After removal from the  
bath, place the unit under gently running water and towel  
dry before storing.  
• Transport your unit cool, dry, and protected.  
a
INSPECTIONSANDSERVICE  
Your Pro Plus 2 should be inspected annually by an  
Authorized Oceanic Dealer who will perform a factory  
prescribed function check and inspection for damage or  
wear. To keep the 2 year limited warranty in effect, this  
inspection must be completed one year after purchase (+/-  
30 days). Oceanic recommends that you continue to have  
this inspection performed every year to ensure it is working  
properly. The costs of annual inspections are not covered  
under the terms of the 2 year limited warranty.  
X
X
X
X
b
Fig.95-CaseBack  
Download from Www.Somanuals.com. All Manuals Search And Download.  
84  
ToObtainService:  
Take your Pro Plus 2 to an Authorized Oceanic Dealer or  
send it to the nearest Oceanic Regional Distributor Facility.  
To return your Pro Plus 2 to Oceanic:  
• Record all dive data in the Log and/or download the  
data in memory. All data will be erased when it receives  
factory service.  
• Package it using a protective cushioning material.  
• Include a legible note stating specific reason for return,  
your name, address, daytime phone number, serial  
number, and a copy of your original sales receipt and  
Warranty Registration Card.  
• Send freight prepaid and insured using a traceable  
method to the nearest Oceanic Regional Service Facility  
(see page 96), or to Oceanic USA.  
• If shipping to Oceanic USA, obtain an RA (Return Autho-  
rization) number by contacting Oceanic at 510/562-  
0500 or send an e-mail to [email protected].  
• Non-warranty service must be prepaid. COD is not  
accepted.  
• Additional information is available at the Oceanic web  
site OceanicWorldWide.com  
E
Download from Www.Somanuals.com. All Manuals Search And Download.  
85  
NOTE: Theproceduresthatfollowmustbeclosely  
adheredto. DamageduetoimproperBatteryreplace-  
mentisnotcoveredbytheunit's2yearwarranty.  
BATTERYREPLACEMENT  
The Battery Compartment should only be opened in a dry  
and clean environment with extreme care taken to prevent  
the entrance of moisture or dust.  
As an additional precautionary measure to prevent formation  
of moisture in the Battery Compartment, it is recommended  
that the Battery be changed in an environment equivalent to  
the local outdoor temperature and humidity (e.g., do not  
change the Battery in an air conditioned environment then  
take it outside during a hot sunny day).  
Inspect the Buttons, Lens, and Housing to ensure they are not  
cracked or damaged. If there is any sign of moisture in the  
module, DO NOT use the Pro Plus 2 until it receives proper  
service by an Authorized Oceanic Dealer, or the Oceanic  
factory.  
HotSwap  
If the new Battery can be inserted within 8 seconds after the  
old one is removed (Hot Swap), settings and nitrogen and  
oxygen calculations for repetitive dives will be retained.  
E
Download from Www.Somanuals.com. All Manuals Search And Download.  
86  
BatteryRemoval  
Examine the Case Back to locate the Battery Hatch (Fig. 96a):  
• Remove the 4 screws that secure the Battery Hatch to the  
Housing by turning counter clockwise.  
• Lift the Hatch up and away from the Housing.  
• Lift the Battery, positive (+) end first, out of the Battery  
Compartment.  
• Remove the Battery Hatch O-ring and inspect it for any  
signs of deterioration or deformity. DO NOT use tools to  
remove the O-ring.  
• O-ring replacement is highly recommended to ensure  
proper sealing.  
• Closely examine the sealing surfaces of the Battery Hatch  
and Housing for any signs of damage that might prevent  
proper sealing. If found, return the Pro Plus 2 to an  
Authorized Oceanic Dealer, and DO NOT attempt to use  
it until it has received factory service.  
• Closely examine the inside of the Battery Compartment for  
any signs of corrosion indicating entrance of moisture into  
the unit.  
X
X
• If corrosion is found, return the Pro Plus 2 to an Autho-  
rized Oceanic Dealer, and DO NOT attempt to use it until  
it has received factory service.  
X
X
a
(continued on page 88)  
Fig.96-BatteryHatch  
Download from Www.Somanuals.com. All Manuals Search And Download.  
87  
MoistureintheBatteryCompartment  
• If moisture is found, it is best to have the unit inspected  
and cleaned by an Authorized Oceanic Dealer.  
• If it is necessary to clean the Battery Compartment, flush  
the it and all components with a solution of 50% white  
vinegar and 50% fresh water. Rinse with fresh water,  
and allow to dry overnight, or blow dry with a hair dryer  
(set at 'no heat').  
+
• Closely check all of the sealing surfaces for any signs of  
damage that might impair proper sealing.  
• Inspect the Buttons, Lens, and Housing to ensure they are  
not cracked or damaged.  
a
Fig.97 -O-ringInstallation  
BatteryInstallation  
• Replace the Hatch O-ring with a new one. This O-ring  
must be a genuine Oceanic part that can be purchased  
from an Authorized Oceanic Dealer. Use of any other  
O-ring will void the warranty.  
• Lightly lubricate the new Hatch O-ring with silicone  
grease and place it on the beveled outer edge of the  
Battery Compartment (Fig. 97a). Ensure that it is evenly  
seated.  
+
• Place a new 3 volt, CR2, .75 AHr, Lithium Battery  
(Duracell model DL-CR2 or equivalent) negative end first  
into the Battery Compartment with the negative end  
toward the spring (Fig. 98a).  
a
Fig.98 -BatteryOrientation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
88  
BatteryHatchInstallation  
• Ensure that the Battery is properly oriented and the Hatch  
O-ring is evenly seated.  
• Carefully place the Battery Hatch into position so that it  
seats on top of the O-ring, and while holding it in place,  
secure it with the 4 screws by turning them clockwise.  
DO NOT attempt to use any other screws.  
• Carefully tighten the screws by alternately turning them  
each one turn at a time (Fig. 99). Turn the upper left one  
(a), then the lower right (b), then the lower left (c), then  
the upper right (d).  
• Repeat the sequence until all of the screws are evenly  
secure. The outer surface of the Battery Hatch should be  
flush with the outer surface of the Housing. DO NOT  
overtighten.  
Inspection  
• Activate the unit and watch carefully as it performs a full  
diagnostic and battery check, and enters Surface Mode.  
Observe the LCD display to ensure it is consistently clear  
and sharp in contrast throughout the screen.  
X
X
a
c
d
b
X
X
• If any portions of the display are missing or appear dim,  
or if a Low Battery Condition is indicated, return your Pro  
Plus 2 to an Authorized Oceanic Dealer for a complete  
evaluation before attempting to use it.  
Fig.99 -HatchInstallation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
89  
SPECIFICATIONS  
CAN BE USED AS  
OPERATIONALMODES(SURFACE)  
AirComputer  
NitroxComputer  
DigitalDepthGauge/Timer  
Activation/Diagnostic  
Surface  
PreDivePlanningSequence  
(30-190feet/9-57meters)  
NODECOMPRESSIONMODEL  
Basis:  
TimetoFlyCountdown  
DesaturationCountdown  
DiveLog(Date/Time,Nitrogen,andOxygen)  
SetMode1:  
ModifiedHaldaneanAlgorithm  
12tissuecompartments  
• FO2(Air,21to50%O2)  
• MaxDepthAlarm(30to300feet/9to99meters)  
• ElapsedDiveTimeAlarm(0:10-3:00hr:min)  
• PCInterface(toDownloaddata)  
SetMode2:  
DataBase:  
DivingScienceandTechnology(DSAT)-Rogers/Powell  
Performance:  
• UnitsofMeasure(Imperial/Metric)  
• HourFormat(12/24)  
• Time(Hour,Minute)  
Tissuecompartmenthalftimes(mins.)Spencer’sMvalues  
5,10,20,40,80,120,160,200,240,320,400,480  
Reciprocalsubsurfaceelimination  
60minutesurfacecreditcontrolforcompartmentsfaster  
than60minutes  
• Date(Year,Month,Day)  
• AudibleAlarm/LEDWarning(On/Off)  
• MaxTLBGAlarm(1to12segments)  
• AirTimeRemainingBGAlarm(0to60minutes)  
• TurnPressureAlarm(1000to3000PSI/69to205BAR)  
• EndPressureAlarm(300to1500PSI/20to104BAR)  
• DiveTimeRemainingAlarm(0:00-0:20min)  
• MaxPO2Alarm(1.20-1.60ATA)  
• FO250%Default(On/Off)  
Tissuecompartmentstrackedupto24hoursafterlastdive  
DecompressionCapabilities(stopceilings):  
10,20,30,40,50,and60feet  
(3,6,9,12,15,and18meters)  
AltitudeAlgorithm:  
BasedonNOAAtables  
• BacklightDuration(:00/:03/:07seconds)  
• SamplingRate-memorystoragefordownload  
(2,15,30,60seconds;2,5,10feet,.5,1.5,3meters)  
• DigitalGaugeMode(On/Off)  
• WaterActivation(On/Off)  
OxygenExposureLimits:  
BasedonNOAAtables  
SimulatorMode  
Download from Www.Somanuals.com. All Manuals Search And Download.  
90  
SPECIFICATIONS(CONTINUED)  
OPERATIONALMODES(DIVE)  
NoDecompressionDive:  
#1-CurrentDepth,DiveTimeRemaining,CylinderPressure,BarGraphs  
#2-CurrentDepth,DiveTimeRemaining,MaxDepth,ElapsedDiveTime,CylinderPressure,BarGraphs  
#3-CurrentDepth,DiveTimeRemaining,Temperature,TimeofDay,CylinderPressure,BarGraphs  
#4-ifanitroxdive -CurrentDepth,DiveTimeRemaining,CurrentPO2,CylinderPressure,BarGraphs  
Secondary-CurrentDepth,AirTimeRemaining,CylinderPressure,BarGraphs  
SafetyStop-CurrentDepth,StopDepth(15FT/5M),CountdownTimer(3:00to0:00min:sec),DiveTimeRemaining,  
CylinderPressure,BarGraphs  
DigitalGaugeMode:  
Default-CurrentDepth,MaxDepth,TimeofDay,ElapsedDiveTime,CylinderPressure,BarGraphs  
Secondary-CurrentDepth,AirTimeRemaining,CylinderPressure,BarGraphs  
DecompressionDive:  
#1-Default-CurrentDepth,StopDepth/Time,TotalAscentTime,CylinderPressure,BarGraphs  
#2-CurrentDepth,StopDepth/Time,TotalAscentTime,MaxDepth,ElapsedDiveTime, CylinderPressure,  
BarGraphs  
#3-CurrentDepth,Temperature,TimeofDay,TotalAscentTime,CylinderPressure,BarGraphs  
#4-ifnitroxdive-CurrentDepth,CurrentPO2,TotalAscentTime,CylinderPressure,BarGraphs  
Violation-Conditional,Delayed,and Immediate/Gauge  
HighPO2(1.20to1.60ATA)  
HighOxygenAccumulation(300OTUperdive/24hr)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
91  
SPECIFICATIONS(CONTINUED)  
DISPLAYRANGE/RESOLUTION  
NumericDisplays:  
Range:  
0 to 24  
0 to 330ft(0to99.9m)  
330ft(99.9m)  
Air, 21to50%  
Resolution:  
1
1ft(.1m)  
1ft(.1m)  
1%  
.01ATA  
1minute  
1minute  
1minute  
DiveNumber  
Depth  
MaximumDepth  
FO2SetPoint  
PO2Value  
DiveTimeRemaining  
AirTimeRemaining  
TotalAscentTime  
Safety  
DecompressionStopTime  
ElapsedDiveTime  
SurfaceTime  
DiveLogSurfaceInterval  
TimetoFly  
0.00to5.00ATA  
0:00to9:59hr:min  
0:00to9:59hr:min  
0:00to9:59hr:min  
Time  
0:00to9:59hr:min  
0:00-9:59hr:min  
0:00-23:59hr:min  
0:00-23:59hr:min  
23:50-0:00hr:min*  
(*starting10minafterthedive)  
23:50-0:00hr:min*  
(*starting10min.afterthedive)  
0 to 99°F(-9to60°C)  
0 to 5000psi(0to352BAR)  
Stop  
3:00  
to  
0:00  
mi  
1minute  
1minute  
1minute  
1minute  
1minute  
TimetoDesaturate  
1minute  
Temperature  
CylinderPressure  
1°  
10psi(.5BAR)  
SpecialDisplays:  
Occurrence  
DiagnosticDisplay  
OutofRange(---)  
GaugeModeCountdownTimer  
AfterManualActivation  
>330feet(>99.9meters)  
23:50to0:00hr:min(afterviolation)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
92  
SPECIFICATIONS(CONTINUED)  
BARGRAPHS  
TissueLoadingBarGraph:  
segments  
Oxygen(O2)BarGraph:  
segments  
NoDecompressionzone(green)  
NoDecoCautionzone(yellow)  
DecompressionWarningzone(red)  
9
2
1
Normalzone(green)  
Cautionzone(yellow)  
Dangerzone(red)  
4
1
1
VariableAscentRateIndicator: 60feet(18m)&Shallower  
segments feet/min meters/min  
Deeperthan60feet(18m)  
segments feet/min meters/min  
0
1
2
3
4
5
0 - 10  
0 - 3  
3.5-4.5  
5 - 6  
6.5-7.5  
8 - 9  
> 9  
0
1
2
3
4
5
0 - 20  
0 - 6  
NormalZone(Green)  
NormalZone(Green)  
NormalZone(Green)  
CautionZone(Yellow)  
TooFastZone(Red-flashing)  
11 - 15  
16 - 20  
21 - 25  
26 - 30  
>30  
21-30  
31 - 40  
41-50  
51 - 60  
>60  
6.5-9  
9.5-12  
12.5-15  
15.5-18  
>18  
OPERATIONALPERFORMANCE  
Function:  
Accuracy:  
Depth  
Timers  
±1%offullscale  
1secondperday  
DiveCounter:  
DisplaysDives#1to24,0ifnodivemadeyet  
ResetstoDive#1,upondiving(newactivationperiod)  
DiveLogMode:  
Stores24mostrecentdivesinmemoryforviewing  
After24dives,adds25thdiveinmemoryanddeletesthefirstdive  
Download from Www.Somanuals.com. All Manuals Search And Download.  
93  
SPECIFICATIONS(CONTINUED)  
OPERATIONALPERFORMANCE(continued)  
Altitude:  
Operationalfromsealevelto14,000feet(4,267meters)elevation  
Measuresambientpressureevery30minutesandwhenmanuallyactivated(nowhenwetcontactsarebridged)  
CompensatesforAltitudewhenmanuallyactivated(nocompensationifactivatedbyimmersioninwater)  
Compensationbeginsat2,000feet(610meters)elevationandevery1,000feet(305meters)higher  
Power:  
• Battery  
• Shelflife  
1-3vdc,CR2,.75Ahr,Lithiumbattery(DuracellmodelDL-CR2orequivalent)  
Upto5years  
• Replacement  
• Lifeexpectancy  
Userreplaceable(annualrecommended)  
50divehours(if1-1hourdiveperdiveday)toover150divehours(if3-1hourdivesperdiveday)  
Battery Indicator:  
segmentsdisplayed  
all  
estimatedpowerremaining  
25to100%  
1(inside)  
<25%  
Activation:  
Manual-pushbutton(recommended)  
Automatic-byimmersioninwater(asabackupifsetON)  
H2OgraphicindicatesWetContactsarebridged(unitmustbedriedpriortotransportorstorage)  
Cannotbemanuallyactivateddeeperthan4feet(1.2meters),iftheWaterActivationfeatureissetOFF.  
Cannotbeactivatedatelevationshigherthan14,000feet(4,267meters)  
Shutoff:  
Automaticallyshutsoffifnodiveismadewithin120minutesafterinitialactivation. Reactivationrequired.  
Automaticallyshutsoff24hoursafterlastdive(willreactivateiftheH2Ographicisdisplayed).  
Cannotbeshutoffmanually.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
94  
SPECIFICATIONS(CONTINUED)  
OPERATIONALPERFORMANCE(continued)  
SettingFO2:  
Automaticallysetfor'Air'uponactivation  
RemainssetforAirunlessanFO2numericalvalueisset  
Nitroxsetpointsfrom21to50%  
Ifsetfor21%,remainssetfor21%untilchanged  
Ifsetfor>21%,itrevertsto50%10minutesafterthedive,iftheFO2DefaultisON. IftheFO2DefaultisOFF,thevaluewill  
remainatthevalueset.  
OperatingTemperature:  
Outofthewater-between20°Fand140°F(-6and60°C)  
Inthewater-between28°Fand95°F(-2and60°C).  
Atextremelylowtemperatures,theLCDmaybecomesluggish,butthiswillnotaffectit'saccuracy. Ifstoredortransportedin  
extremelylowtemperatureareas(belowfreezing),youshouldwarmtheunitanditsbatterywithbodyheatbeforediving.  
ACCESSORIES(optionalitemsavailablefromyourAuthorizedOceanicDealer):  
• LensProtector -adherestolensface,preventsscratches  
• PCdownloadpackage(hardwareandsoftware)  
• BatteryKit-includes1Battery,1BatteryHatchO-ring,SiliconeGrease  
Download from Www.Somanuals.com. All Manuals Search And Download.  
95  
OCEANIC WORLD WIDE  
OCEANIC USA  
2002 Davis Street  
San Leandro, CA 94577  
Tel: 510/562-0500  
Fax: 510/569-5404  
Web site: http://www.OceanicWorldWide.com  
Oceanic Germany - Nurnberg, Germany  
Tel:49-911-324-6630 Fax:49-911-312-999  
OceanicDivingAustraliaPty.Ltd  
Sorrento,Victoria,Australia  
Tel:61-3-5984-4770 Fax:61-3-5984-4307  
OceanicSouthEurope - Genova, Italy  
Tel:0039-010-834-51 Fax:0039-010-834-52-50  
Oceanic Asia-Pacific Pte. Ltd - Singapore  
Tel: 65-779-3853 Fax: 65-779-3945  
Oceanic SW, Ltd - Devon, United Kingdom  
Tel:44-1-404-89-1819 Fax:44-1-404-89-1909  
OceanicJapan - Yokohama,Japan  
Tel: 045-575-6671 Fax: 045-575-6673  
OceanicFrance - Marseille,France  
Tel: 33-491-25-27-45 Fax: 33-491-25-35-86  
OceanicNewZealand  
Wellington,NewZealand  
Tel: 64-4-472-5335 Fax: 64-4-472-5334  
OceanicInternational(Pacific) - Kapolei,Hawaii  
Tel: 808-682-5488 Fax: 808-682-1068  
Download from Www.Somanuals.com. All Manuals Search And Download.  
96  
INSPECTION / SERVICE RECORD  
SerialNumber ________________________  
Dateofpurchase _____________________  
Purchasedfrom ______________________  
BelowtobefilledinbyanAuthorizedOceanicDealer:  
E
Date  
ServicePerformed  
Dealer / Technician  
Download from Www.Somanuals.com. All Manuals Search And Download.  
97  
NOTES  
Download from Www.Somanuals.com. All Manuals Search And Download.  
98  
RESET FEATURE  
This dive computer is configured with a RESET feature that allows data to be cleared,  
including Nitrogen and Oxygen calculations, FO2 set point, Log Mode entries, and  
Download data.  
WARNING: Resetafteradiveandsubsequentuseforarepetitivediveconducted  
bythesamedivercouldresultinseriousinjurytoordeath.  
RESET PROCEDURE  
• While in SURFACE MODE (new activation period or greater than a 10 minute post  
dive surface interval), press the Left (Advance) button one (1) time to access PLAN  
MODE.  
• While 30 feet (or 9 meters) is displayed in the PLAN MODE, press and hold Both  
buttons until SET 2 appears, then release the buttons.  
• Press and release the Left (Advance) button to display the first 2 digits of the assigned  
KEY CODE, flashing as 00.  
• Press and release the Right (Select) button to increase the digits to 01.  
• Press and release the Left (Advance) button again to display the second 2 digits of the  
KEY CODE, flashing as 00.  
• Press and release the Right (Select) button to increase the digits to 01.  
• Press and release the Left (Advance) button to complete the RESET operation and turn  
the unit OFF.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
99  
®
OCEANIC USA  
2002 Davis Street  
San Leandro, CA 94577  
Tel: 510-562-0500  
Fax: 510-569-5404  
http://www.OceanicWorldWide.com  
© 2002 Design 2002  
Doc. No. 12-2328-r03 (6/6/02)  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Milwaukee Drill 0100 20 User Manual
Nady Systems Microphone CM 60 User Manual
NorthStar Navigation GPS Receiver 941X User Manual
Nvidia Computer Hardware DG 04927 001_V01 User Manual
Oki Printer 6120E User Manual
Oki Printer Accessories MB260 User Manual
Optimus Home Theater System STAV 3580 User Manual
P3 International Pest Control Equipment P7900
Panasonic Electric Shaver ER206 User Manual
Panasonic Projector D10000 User Manual