National Instruments Switch GPIB 100A User Manual

GPIB-100A  
User Manual  
March 1990 Edition  
Part Number 320063-01  
© Copyright 1985, 1991 National Instruments Corporation.  
All Rights Reserved.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Limited Warranty  
The GPIB-100A is warranted against defects in materials and workmanship for a period of two years from the  
date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option,  
repair or replace equipment that proves to be defective during the warranty period. This warranty includes  
parts and labor.  
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the  
outside of the package before any equipment will be accepted for warranty work. National Instruments will  
pay the shipping costs of returning to the owner parts which are covered by warranty.  
National Instruments believes that the information in this manual is accurate. The document has been  
carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National  
Instruments reserves the right to make changes to subsequent editions of this document without prior notice to  
holders of this edition. The reader should consult National Instruments if errors are suspected. In no event  
shall National Instruments be liable for any damages arising out of or related to this document or the  
information contained in it.  
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR  
IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A  
PARTICULAR PURPOSE. CUSTOMER'S RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE  
ON THE PART OF NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY  
THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS  
OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED  
OF THE POSSIBILITY THEREOF. This limitation of the liability of National Instruments will apply regardless of  
the form of action, whether in contract or tort, including negligence. Any action against National Instrument  
must be brought within one year after the cause of action accrues. National Instruments shall not be liable for  
any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not  
cover damages, defects, malfunctions, or service failures caused by owner's failure to follow the National  
Instruments installation, operation, or maintenance instructions; owner's modification of the product; owner's  
abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or  
other events outside reasonable control.  
Copyright  
Under the copyright laws, this book may not be copied, photocopied, reproduced, or translated, in whole or in  
part, without the prior written consent of National Instruments Corporation.  
Trademarks  
Product names listed are trademarks of their respective manufacturers. Company names listed are trademarks  
or trade names of their respective companies.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
FCC/DOC Radio Frequency Interference Compliance  
This equipment generates and uses radio frequency energy and, if not installed and used in strict  
accordance with the instructions in this manual, may cause interference to radio and television  
reception. This equipment has been tested and found to comply with (1) the limits for a Class A  
computing device, in accordance with the specifications in Subpart J of Part 15 of U.S. Federal  
Communications Commission (FCC) Rules, and (2) the limits for radio noise emissions from  
digital apparatus set out in the Radio Interference Regulations of the Canadian Department of  
Communication (DOC). These regulations are designed to provide reasonable protection against  
interference from the equipment to radio and television reception in commercial areas.  
There is no guarantee that interference will not occur in a particular installation. However, the  
chances of interference are much less if the equipment is used according to this instruction manual.  
If the equipment does cause interference to radio or television reception, which can be determined  
by turning the equipment on and off, one or more of the following suggestions may reduce or  
eliminate the problem.  
Operate the equipment and the receiver on different branches of your AC electrical system.  
Move the equipment away from the receiver with which it is interfering.  
Relocate the equipment with respect to the receiver.  
Reorient the receiver's antenna.  
Be sure that the equipment is plugged into a grounded outlet and that the grounding has not  
been defeated with a cheater plug.  
If necessary, consult National Instruments or an experienced radio/television technician for  
additional suggestions. The following booklet prepared by the FCC may also be helpful: How to  
Identify and Resolve Radio-TV Interference Problems. This booklet is available from the U.S.  
Government Printing Office, Washington, DC 20402, Stock Number 004-000-00345-4.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Preface  
Organization of the Manual  
This manual is divided into the following chapters:  
Chapter 1, Description of the GPIB-100A, contains a brief description of the GPIB-100A Bus  
Extender and how it is used. This section also lists all components and accessories. In addition, it  
provides system configuration, performance, operating, electrical, environmental, and physical  
specifications for the GPIB-100A.  
Chapter 2, Installation, contains instructions for configuring and connecting the GPIB-100A into  
your system at your operating voltage.  
Chapter 3, Configuration and Operation, describes how to configure and operate a GPIB-100A  
system.  
Chapter 4, Theory of Operation, contains descriptions of how the GPIB-100A circuitry operates.  
Appendix A, Operation of the GPIB, describes GPIB terminology and protocol for users  
unfamiliar with the GPIB.  
Appendix B, Schematic Diagram, contains a detailed schematic diagram of the GPIB-100A.  
Appendix C, GPIB-100A Parts Locator Diagram, contains the parts locator diagram for the  
GPIB-100A.  
Appendix D, Cable Assembly Wire List, contains the listing of wire connections for the  
GPIB-100A transmission cable.  
Appendix E, Multiline Interface Messages, contains an ASCII chart and a list of the corresponding  
GPIB messages.  
Appendix F, Mnemonics Key, contains a mnemonics key that defines the mnemonics used  
throughout the manual.  
Related Document  
The following document is a reference that covers in greater detail specific topics introduced in this  
manual:  
ANSI/IEEE Standard 488-1978, IEEE Standard Digital Interface for Programmable  
Instrumentation.  
© National Instruments Corporation  
v
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Preface  
Abbreviations Used in the Manual  
The following abbreviations are used in the text of this manual.  
C
F
centigrade  
Fahrenheit  
Hz  
hertz  
in.  
inch  
kbytes  
m
thousand bytes  
meter  
mA  
Mbytes  
mm  
µsec  
nsec  
sec  
milliamperes  
million bytes  
millimeter  
microsecond  
nanosecond  
second  
V
volts  
VAC  
W
Volts Alternating Current  
watt  
GPIB-100A User Manual  
vi  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Chapter 1  
Introduction...................................................................................................................... 1-1  
Chapter 2  
Inspection......................................................................................................................... 2-1  
Power Connection............................................................................................................ 2-1  
Grounding Configuration................................................................................................. 2-2  
Disassembly ........................................................................................................2-2  
Connecting to Hewlett-Packard Controllers..................................................................... 2-2  
Chapter 3  
Operating Modes .............................................................................................................3-1  
Talker/Listener/Controller (TLC) Mode............................................................... 3-1  
Talker/Listener (TL) Mode................................................................................... 3-2  
Parallel Poll Response (PPR) Modes............................................................................... 3-2  
Buffered PPR Mode (Approach 1)...................................................................... 3-3  
Mixed Mode Operation.................................................................................................... 3-4  
Chapter 4  
Diagrams ......................................................................................................................... 4-1  
Power-On ........................................................................................................................ 4-1  
System Controller Detection............................................................................................ 4-2  
Source Handshake Detection ........................................................................................... 4-2  
Data Direction Control......................................................................................... 4-4  
© National Instruments Corporation  
vii  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Appendix A  
Operation of the GPIB....................................................................................................... A-1  
Types of Messages .......................................................................................................... A-1  
Talkers, Listeners, and Controllers...................................................................................A-1  
Data Lines........................................................................................................................ A-3  
NRFD (not ready for data)................................................................................... A-3  
NDAC (not data accepted)................................................................................... A-4  
DAV (data valid) ................................................................................................. A-4  
Interface Management Lines............................................................................................ A-4  
ATN (attention).................................................................................................... A-4  
REN (remote enable)........................................................................................... A-4  
SRQ (service request).......................................................................................... A-4  
EOI (end or identify) ........................................................................................... A-4  
Appendix B  
Appendix C  
Appendix D  
Cable Assembly Wire List...............................................................................................D-1  
Multiline Interface Messages......................................................................................... E-1  
Appendix F  
GPIB-100A User Manual  
viii  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Figures  
Figure 1-1. The Model GPIB-100A Bus Extender.................................................................. 1-1  
Figure 1-2. Typical GPIB-100A Extension System (Physical Configuration) ........................ 1-2  
Tables  
Table 1-1. System Configuration Characteristics......................................................................1-3  
Table 1-6. Physical Characteristics............................................................................................1-6  
© National Instruments Corporation  
ix  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 1  
Description of the GPIB-100A  
Introduction  
The high-speed GPIB-100A Bus Extender (Figure 1-1) is used in pairs with a special parallel data  
transmission cable to connect two separate GPIB or IEEE-488 bus systems in a functionally  
transparent manner.  
Figure 1-1. The Model GPIB-100A Bus Extender  
While the two bus systems are physically separate, as shown in Figure 1-2, devices logically  
appear to be located on the same bus as shown in Figure 1-3. Thus, with the GPIB-100A it is  
possible to overcome two configuration restrictions imposed by ANSI/IEEE Standard 488-l978,  
namely:  
Cable length limit of 20 m total per contiguous bus or 2 m times the number of devices on the  
bus, whichever is smaller.  
Electrical loading limit of 15 devices per contiguous bus.  
Each GPIB-100A system extends the distance limit by 300 m and the loading limit to 30 devices  
including the extenders, without sacrificing speed or performance. These point-to-point extender  
systems can be connected in series for longer distances or in star patterns for additional loading.  
At short distances, the data transfer rate over the extension can exceed 250 kbytes/sec, degrading  
with distance only by the propagation delay along the cable. Furthermore, regardless of the  
distance, there is no speed degradation at all for transfers between devices on the same side of the  
© National Instruments Corporation  
1-1  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Description of the GPIB-100A  
Chapter 1  
extension. And because the GPIB-100A is a functionally transparent extender, the same GPIB  
communications and control programs that work with an unextended system will work  
unmodified with an extended system. There is one minor exception to this transparency in  
conducting parallel polls, as explained in Chapter 3 in the paragraph Parallel Poll Response (PPR)  
Modes.  
RS-232 Compatible  
Transmission  
Lines  
GPIB #2  
GPIB #1  
GPIB-100A  
GPIB-100A  
Computer  
(System Controller,  
Talker, and Listener)  
Printer  
(Listener)  
Multimeter  
(Talker and Listener)  
Signal Generator  
(Listener)  
Unit Under Test  
Figure 1-2. Typical GPIB-100A Extension System (Physical Configuration)  
GPIB  
Computer  
(System Controller,  
Talker, and Listener)  
Printer  
(Listener)  
Multimeter  
(Talker and Listener)  
Signal Generator  
(Listener)  
Unit Under Test  
Figure 1-3. Typical GPIB-100A Extension System (Logical Configuration)  
GPIB-100A User Manual  
1-2  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 1  
Description of the GPIB-100A  
GPIB-100A Specifications  
The following tables show the system configuration; the performance, operating, electrical,  
environmental, and physical characteristics of the GPIB-100A, as well as providing a list of  
available GPIB-100A components and accessories.  
Table 1-1. System Configuration Characteristics  
Characteristic  
Specification  
distance per extension  
loading per extension  
multiple extensions  
up to 300 m  
up to 14 additional devices  
permitted in any combination of star or linear pattern  
GPIB driver output  
circuit and T1 timing  
of source device  
no restrictions (automatic conversion to 2 µsec  
T1 delay on remote side is built in)  
Note: T1 is the data settling time (DIO valid to DAV) and varies according to the type of  
drivers and the system configuration used.  
© National Instruments Corporation  
1-3  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Description of the GPIB-100A  
Chapter 1  
Table 1-2. Performance Characteristics  
Characteristic  
Specification  
speed  
250 to 135 kbytes/sec (approximately 4 µsec per byte  
degraded at 10.5 nsec per meter of distance) no  
limitation to device speeds when there are no  
on remote side of extension  
listeners  
functionality  
transparent GPIB operation except for pulsed parallel  
polls  
interlocked  
IEEE-488 handshake  
maintained across the extension  
(message-grams not used)  
IEEE-488 capability  
identification  
codes  
SH1  
complete Source Handshake  
complete Acceptor Handshake  
complete Talker  
AH1  
T5,TE5  
L3,LE3  
SR1  
complete Listener  
complete Service Request  
complete Remote Local  
complete Parallel Poll  
complete Device Clear  
complete Device Trigger  
complete Controller  
RL1  
PP1,2  
DC1  
DT1  
C1-5  
E1  
open collector GPIB drivers  
Table 1-3. Operating Characteristics  
Characteristic  
Specification  
architecture  
asynchronous (no clock) parallel design  
point-to-point (not multi-drop) transmission  
operating modes  
Talker/Listener/Controller or  
Talker/Listener (Talk Only)  
Parallel Poll Response  
modes  
Buffered Parallel Poll Response or  
Unbuffered Parallel Poll Response  
GPIB-100A User Manual  
1-4  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 1  
Description of the GPIB-100A  
Table 1-4. Electrical Characteristics  
Characteristic  
Specification  
GPIB interface circuit  
duplex transceivers with open  
collector drivers (MC3441A)  
transmission interface  
circuit  
RS-422 drivers and receivers  
(MC3487 and AM26LS32) connected with patented  
data transmission cable for minimum skewing  
(<= 3%) between any two pairs  
power supply  
selectable (fuse)  
50 to 60 Hz  
110 V, 160 mA (250 mA, 250 V, Slow Blow)  
220 V, 80 mA (200 mA, 250 V, Slow Blow)  
GPIB interface load  
power  
one standard load, AC and DC  
18 W typical  
Table 1-5. Environmental Characteristics  
Characteristic  
operating temperature  
humidity  
Specification  
0 to 55 C  
5 to 95% non-condensing conditions  
Class A verified  
UL Listed  
FCC  
110V Version  
220V Version  
UL Listed and also classified by Underwriters  
Laboratories Inc. in accordance with International  
Electrotechnical Commission publication 950  
© National Instruments Corporation  
1-5  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Description of the GPIB-100A  
Chapter 1  
Table 1-6. Physical Characteristics  
Characteristic  
case style  
size  
Specification  
CS2  
3.5 x 8.5 x 13 in. (89 x 216 x 330 mm)  
case material  
UL94V-0 flame retardant polystyrene  
Dow 60875 F or equivalent  
rack mounting  
GPIB cable  
single or dual kits available  
Hewlett Packard 10833 style or equivalent  
Dynatronics D-200-24 cable with  
Transmission cable  
AMP Amplimite connectors AMP HDP-20 50 pin  
connector with RFI/EMI shield  
Table 1-7. Components and Accessories  
Item  
Part Number  
Model GPIB-100A Bus Extender (110V)  
(two required per extension)  
776107-01  
Model GPIB-100A Bus Extender (220V)  
(two required per extension)  
776107-31  
Type T2 Transmission Cable  
178056-xxx  
(xxx = length in meters)  
Type X2 GPIB Cable  
1 meter  
2 meters  
763061-01  
763061-02  
763061-03  
4 meters  
Single Rack-Mount Kit  
Dual Rack-Mount Kit  
180304-01  
180304-02  
Note: All part numbers in this table are National Instruments part numbers.  
GPIB-100A User Manual  
1-6  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 2  
Installation  
Inspection  
Inspect the shipping container and contents for evidence of physical damage or stress. If damage  
is discovered and appears to have been caused in shipment, file a claim with the carrier. If the  
equipment is damaged, do not attempt to operate it before contacting National Instruments for  
instructions. Retain the shipping material for possible inspection by carrier or reshipment of the  
equipment.  
Power Connection  
The GPIB-100A Bus Extender is shipped from the factory set at a certain operating voltage, either  
110 VAC or 220 VAC. Verify that the voltage you are using is the same as that selected on the  
rear panel of the GPIB-100A. Operating at a voltage other than the one selected may damage the  
unit. If the GPIB-100A is set at a voltage other than the one you are using, follow the steps below  
to change the operating voltage.  
1. Remove the power cord from the unit.  
2. Pull out the fuse holder and replace the fuse with one that has the type and rating specified in  
Table 1-4 for your operating voltage.  
3. Using a small flat-head screwdriver, rotate the voltage selector to point to your operating  
voltage.  
Figure 2-1. Voltage Selection  
© National Instruments Corporation  
2-1  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Installation  
Chapter 2  
Grounding Configuration  
A U.S. standard three-wire power cable is provided with the GPIB-100A. When connected to a  
power source, this cable connects the equipment chassis to the power ground.  
The GPIB-100A is shipped from the factory with chassis and power grounds connected to the  
logic ground of the digital circuitry and the shields of the interfacing cables. If it is necessary to  
isolate these grounds to prevent current loops between units, disassemble the unit according to the  
following instructions and remove jumper W1 located on the circuit card assembly near the back  
panel.  
Disassembly  
The case consists of two identical sections. Before disassembling, remove power from the unit.  
Then remove the two screws on each side of the case and lift the top section. When reassembling,  
it may be necessary to adjust the two trim panels on the case side for proper fit in their grooves.  
Mounting  
The GPIB-100A enclosure is designed for table top operation or for rack mounting. Single and  
dual unit rack mounting kits are available from National Instruments for field installation.  
Connecting to Hewlett-Packard Controllers  
To achieve very high data transfer rates and long cable spans between devices, many Hewlett-  
Packard (HP) controllers and computers, such as the 64000 series, use a preload technique on the  
unit designated Master Controller. When preloaded, the GPIB lines of the Master Controller are  
terminated to represent six device loads. HP has two types of preloading: Class A, in which all 16  
GPIB lines are loaded, and Class B, in which all lines except Not Ready For Data (NRFD) and  
Not Data Accepted (NDAC) are loaded.  
Preloading increases ringing on signal transitions and may cause improper operation of the GPIB-  
100As. If this happens, all signals on the Master Controller should be set to normal (1 unit) load.  
This is done by means of a back panel switch when working from the exterior. In addition, the  
cabling rule of no more than 2m/device must be strictly enforced.  
GPIB-100A User Manual  
2-2  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 3  
Configuration and Operation  
Users who are unfamiliar with the GPIB should first read Appendix A, Operation of the GPIB, to  
become familiar with GPIB terminology and protocol.  
In the following discussions, the terms local and remote refer to certain states of the two GPIB-  
100A Bus Extenders in the system. When one extender is in a local state, meaning that the state in  
question originated on the local state's side, the other extender is in the corresponding remote state.  
The three states in question are the System Controller, Active Controller, and Source Handshake  
states.  
Operating Modes  
The GPIB-100A has two operating modes: Talker/Listener/Controller mode and Talker/Listener  
mode. Both units in the extension system must be set to the same mode.  
Talker/Listener/Controller (TLC) Mode  
The GPIB-100A is set at the factory to the more common TLC operating mode. The TLC mode  
requires a System Controller on one side of the extension. There may be any number of Talkers,  
Listeners, and other Controllers in the system.  
In the TLC mode, the two GPIB-100As expect to see in order: first the Interface Clear (IFC)  
signal from the System Controller; second the Attention (ATN) signal from the Active Controller;  
and third the Data Valid (DAV) signal from the Active Controller or Talker. A brief description of  
this mode is in the following paragraph.  
Both units power up in a quiescent condition with no local or remote state active. They remain that  
way until one unit detects an IFC pulse from the System Controller which is on the same  
contiguous bus. That unit enters the Local System Controller (LSC) state and causes the other unit  
to enter the Remote System Controller (RSC) state. The IFC and Remote Enable (REN) signals  
are switched to flow from the local to the remote unit. Next, one unit detects the ATN signal from  
the Active Controller, enters the Local Active Controller (LAC) state, and places the other unit in  
the Remote Active Controller (RAC) state. The ATN signal is switched to flow from local to  
remote side and the Service Request (SRQ) is switched to flow in the opposite direction. Finally,  
one unit detects the DAV from the Source Handshake function of the Talker or Active Controller.  
That unit enters the Local Source (LS) state and places the other unit in the Remote Source (RS)  
state. The DAV and Data (DIO) signals are switched to flow from local to remote side, and the  
Not Ready for Data (NRFD) and Not Data Accepted (NDAC) signals are switched to flow from  
remote to local side.  
As the source side for these three key signals–IFC, ATN, and DAV–change, the local/remote  
states of each extender and the directions of the other GPIB signals change accordingly. Chapter 4,  
Theory of Operation, contains a more thorough discussion of this.  
© National Instruments Corporation  
3-1  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Configuration and Operation  
Chapter 3  
Talker/Listener (TL) Mode  
There is no Controller and only one Talker in the TL mode of operation, sometimes called talk only  
mode. Usually, there is just one Listener as well. In the TL mode, the System and Active  
Controller states remain inactive and the IFC, REN, ATN, and SRQ signals are unused. The  
directions of the other signals are set the first time the Talker asserts DAV.  
Setting the Operating Mode  
Both GPIB-100As in the extension system must be set to the same operating mode. Use switch  
S1, position 1, on the back panel of each GPIB-100A to set the operating mode. Set the switches  
as shown in Figure 3-1.  
T / L / C  
1
2 3  
1 2 3  
O
N
O
N
T / L  
A. Talker/Listener/Controller Mode  
B. Talker/Listener Mode  
represents the side of the switch you press down  
Figure 3-1. Switch Settings for Operating Mode  
Parallel Poll Response (PPR) Modes  
According to ANSI/IEEE Standard 488-l978, devices must respond to a parallel poll within 200  
nsec after the Identify (IDY) message (Attention (ATN) and End Or Identify (EOI)) is asserted by  
the Active Controller, which then waits until 2 µsec or more to read the Parallel Poll Response  
(PPR). It is not possible for a remote device on an extended system to respond to this quickly  
because of cable propagation delay. GPIB extender manufacturers have approached this in three  
ways:  
Approach 1:  
Approach 2:  
Approach 3:  
Respond to IDY within 200 nsec with the results of the previous poll of the  
remote bus.  
Ignore the 200 nsec rule and assume the Controller will wait sufficiently long to  
capture the response.  
Do not support parallel polling at all.  
The GPIB-100A uses either Approach 1 or 2, selected at switch S1, position 3. Set this switch as  
shown in Figure 3-2.  
GPIB-100A User Manual  
3-2  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 3  
Configuration and Operation  
Buffered PPR Mode (Approach 1)  
Most Controllers pulse the IDY signal for a period of time exceeding 2 µsec and expect a response  
within that time. When used with this type of Controller, the GPIB-100A should be left in the  
Buffered PPR mode as set at the factory.  
In this mode, the local GPIB-100A extender responds to IDY by outputting the contents of the  
PPR data register. At the same time, a parallel poll message is sent to the remote bus and the poll  
response is returned to the local unit in due course. When the local IDY signal is unasserted, the  
register is loaded with the new remote response. Consequently the register contains the response  
of the previous poll. To obtain the response of both local and remote buses, the control program  
executes two parallel polls back-to-back and uses the second response.  
The software driver library of most Controllers contains an easy-to-use parallel poll function. If,  
for example, the function is called PPOLLand the control program is written in BASIC, the  
sequence to conduct a poll in Buffered PPR mode might be like this:  
CALL PPOLL(PPR)  
CALL PPOLL(PPR)  
IF PPR > 0 GOTO NNN  
If two GPIB extender systems are connected in series, three polls are necessary to get responses  
from the local, middle, and far buses.  
Unbuffered PPR Mode (Approach 2)  
Many Hewlett-Packard GPIB Controllers remain in a parallel poll state with IDY asserted  
whenever they are not performing another function. A change in the response causes an interrupt  
of the control program. In other Controllers, the IDY signal is toggled on and off and the duration  
of the signal can be varied to accommodate delayed responses over extenders. When used with  
these types of Controllers, the GPIB-100A should be set to Unbuffered PPR mode. This means  
that the IDY message is sent to the remote bus and the response is returned as fast as propagation  
delays allow. The Controller must allow time to receive the response.  
© National Instruments Corporation  
3-3  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Configuration and Operation  
Chapter 3  
Mixed Mode Operation  
If there are multiple Controllers and all of the same type are located on the same side of the  
extension, the two GPIB-100A units can be set to Unbuffered and Buffered PPR modes  
accordingly.  
BUF  
1 2 3  
1 2 3  
P
P
R
P
P
R
O
N
O
N
UNBUF  
A. Unbuffered PPR Mode  
B. Buffered PPR Mode  
represents the side of the switch you press down  
Figure 3-2. Switch Settings for Parallel Poll Response Mode  
Operating the GPIB-100A System  
The GPIB-100A extension system is fully operational when power is applied to both units. In  
TLC mode, it is sometimes necessary to power on the System Controller last, after the extenders  
and all other devices are operating. This is necessary if the System Controller executes only one  
IFC shortly after power-on.  
The preferred operating mode is to keep both extenders and at least two-thirds of the devices on  
both buses powered on when there is any GPIB activity.  
GPIB-100A User Manual  
3-4  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 4  
Theory of Operation  
Diagrams  
Figure 4-1 shows a block diagram for the GPIB-100A. Refer to Appendix B for GPIB-100A  
schematic diagrams and Appendix C for the GPIB-100A parts locator diagram.  
Figure 4-1. GPIB-100A Block Diagram  
Power-On  
When the GPIB-100A is powered on, a reset pulse (PON) created by U48F, U28A/D and  
associated Register/Capacitor Delay (RCD) network directly or indirectly clears all flip-flops (FFs)  
to an initialized state. PON remains active until both units in the extension are powered on.  
© National Instruments Corporation  
4-1  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
Chapter 4  
System Controller Detection  
PON initializes FFs U22A and U12A to clear the Remote System Controller (RSC) and Local  
System Controller (LSC) signals.  
When Interface Clear (IFC) is received from the local side via GPIB transceiver U2B, the LSC FF  
is set on the leading edge of IFC and after a delay through U21B/C/D/E, IFC is enabled (U35D) to  
the remote unit as XIFC through driver U29A. LSC enables the local unit to transmit Remote  
Enable (REN) to the remote unit through driver U8A.  
XIFC becomes RIFC on the remote side and is received through U10D and propagated to the  
remote GPIB through transceiver U2B, where the bus signal is received back to clock the RSC FF  
U22A. RSC enables REN (U32C) to be driven on the remote GPIB through transceiver U2D.  
Active Controller Detection  
The Remote Active Controller (RAC) and Local Active Controller (LAC) FFs U22B and U12B  
remain cleared until either RSC or LSC is set via U41B–that is, until the System Controller has  
been located. After a short delay (U21A/F and U31B/F), the Attention (ATN) receiver on the local  
side (U32D) is enabled. This delay allows the LAC FF to be set if ATN is already asserted when  
IFC occurs.  
When ATN is received from the local side via transceiver U2C, the LAC FF is set on the leading  
edge of ATN and, after a delay through U11B/C/D/E, ATN is enabled (U32A and U43D) to the  
remote unit as XATN through driver U29B. LAC enables the local unit to receive Service Request  
(SRQ) from the remote unit through receiver U10B, U32B, and transceiver U2A.  
XATN becomes RATN on the remote side and is received through U20B and propagated to the  
remote GPIB through transceiver U2C, where the bus signal is received to clock RAC FF U22B.  
When RAC is set, drivers U8C/D, which transmit SRQ and parallel poll handshake signal BUS  
PP to the local unit, are enabled (that is, toward the Active Controller).  
Source Handshake Detection  
The Local Source (LS) handshake FF U45A is cleared via U33C on the following events:  
Before the Active Controller is identified (TLC mode only)  
Whenever a change in the state of the local ATN signal is caused by a pulse created via U38D,  
U24A/D, and associated RC network.  
While ATN or Data Valid (DAV) is received from the remote unit (U34B).  
During a parallel poll (U46C).  
The Remote Source (RS) handshake FF U36B is cleared via U33B on the following events:  
Before the Active Controller is identified (TLC mode (U33A) only).  
GPIB-100A User Manual  
4-2  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 4  
Theory of Operation  
Whenever a change in the state of the local ATN signal is caused by a pulse created via U38E,  
U24B/C, and associated RC network.  
While ATN or DAV is received from the local side (U34A/D).  
During a parallel poll (U38F and U34A).  
Before the LS FF is set and unless a parallel poll is in progress, the unit drives the local Not Ready  
for Data (NRFD) signal passively false (U42C and U41C). After the Active Controller is  
identified and before the Source Handshake is identified, the unit drives Not Data Accepted  
(NDAC) signals (U42B and U41D) true. Thus, the unit appears in a normal RFD/NDAC state to  
the local GPIB which is awaiting the first data or command byte.  
When DAV is received, it is first delayed slightly by U23A, U48E, and associated RCD network  
and enabled (U25A) to the LS FF. Setting the FF causes the Local Source Handshake to wait until  
ATN changes have propagated and any parallel poll completes fully. The purpose of the DAV  
delay is to filter tail-end unstable transitions from a fast rising edge. DAV is further delayed  
through U23B/C/E/F before being enabled at U25C by LS to be transmitted to the remote side as  
XDAV through driver U30B.  
XDAV is received as RDAV on the remote side through U19D. The signal sets the RS FF after  
all clearing conditions are removed (U35B). DAV is delayed 2 µsec or more through U38C,  
U48B, and associated RCD network to ensure proper data setup time (T1) on the remote side.  
Once RS is set and the remote GPIB is ready for data (U46A), DAV is allowed to propagate  
(U46D and U36C) to the remote GPIB through U47D, U27C, and GPIB transceiver U1B, and  
NRFD is transmitted to the other side through driver U30A (XRFD).  
Once the LS FF is set, the propagation of NRFD from the remote side sets FF U36A via receiver  
U19A, U44C, and U35A. At this point, the unit drives the NRFD and NDAC lines according to  
the levels sensed at the remote unit (via U42C,U41A/D, and GPIB transceiver U1C for NDAC).  
Parallel Polling  
When the local unit detects ATN and End Or Identify (EOI) asserted at the same time, regardless  
of which occurs first, FF U45D is set via U26A, U44E, U48D, and U43C. This causes EOI to be  
transmitted to the remote side as XEOI through U46B and driver U30C. ATN is also transmitted  
to the remote side as XATN through U43D and driver U29B. XEOI and XATN remain asserted  
until the poll signals propagate to the remote unit and a response is returned, even if the local  
signals become unasserted in the meantime. To prevent the local side from further non-poll  
activity, NRFD is asserted via U46C, U41C, and transceiver U1D.  
If the Buffered PPR mode is selected, the contents of the PPR register (U16) are routed through  
the A side of multiplexers U13 and U14 to the local GPIB. The A side is selected whenever the  
local unit is not being polled from the other side (U27B) and the RS FF is cleared (U47A).  
XEOI and XATN are received on the remote side as REOI and RATN through receivers U19B  
and U20B and propagated to the remote GPIB. Two microseconds later, a parallel poll handshake  
signal (U27C, U38A, and associated RCD network) is transmitted back to the local side through  
driver U8C as the signal BUS PP.  
BUS PP is received at the local unit through U10C. When the local poll is over (ATN or EOI  
unasserted), FF U45B is cleared and U36D is set (via U37B/C and U26A). Setting U36D latches  
© National Instruments Corporation  
4-3  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Theory of Operation  
Chapter 4  
the remote poll response into register U16. Clearing U45B unasserts XEOI and XATN, and after  
they propagate to the remote side, BUS PP is also unasserted. This causes FF U36D to be cleared  
as well, terminating the parallel poll process and removing the NRFD condition of the local  
extender.  
To recap, FF U45B is set from the start of the local poll until the remote response is available and  
the local poll is over. FF U36D is set from the time U45B is cleared until the remote poll  
handshake is over. While either is set, the local unit remains in an NRFD holdoff.  
Data Direction Control  
The unit drives the GPIB data lines DI01 to DI08 through transceivers U3 and U4 if there is a  
local parallel poll in progress (U47B and U26A) or if the RS FF is set and a remote parallel poll is  
not in progress (U47A and U27B). Otherwise, these lines are not driven.  
The source for these data lines when they are driven is the remote unit through receivers U15 and  
U17 when Unbuffered PPR mode is selected (Switch S1, position 3 open) or when the RS FF is  
set and a remote parallel poll is not in progress (U47A and U27B). Otherwise, the source is the  
Buffered PPR register U16.  
The unit drives the transmission data lines BUS DIO1-8 through drivers U5 and U7 if there is a  
remote parallel poll in progress (U37D and U27B) or if the LS FF is set and a local parallel poll is  
not in progress (U47C and U26A). Otherwise, these lines are not driven.  
EOI  
The local unit transmits EOI to the remote side as XEOI if the LS FF is set (transceiver U1A,  
U25B, U46B,and driver U30C). Furthermore, XEOI is asserted from the start of a local parallel  
poll until the poll handshake signal BUS PP is received from the remote unit and the local poll  
stops.  
XEOI is received as REOI at the remote unit through receiver U19B. It propagates to the remote  
GPIB if the local unit is conducting a parallel poll (U27A/B, U37A and transceiver U1A) or if the  
RS remote response (RR) is set and the local unit is not conducting a poll (U47A).  
GPIB-100A User Manual  
4-4  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix A  
Operation of the GPIB  
History of the GPIB  
The GPIB is a link, bus, or interface system through which interconnected electronic devices  
communicate. Hewlett-Packard invented the GPIB, which they call the HP-IB, to connect and  
control programmable instruments manufactured by them. Because of its high system data rate  
ceilings of from 250 kbytes/sec to 1 Mbytes/sec per second, the GPIB quickly became popular in  
other applications such as intercomputer communication and peripheral control. It was later  
accepted as the industry standard IEEE-488. The versatility of the system prompted the name  
General Purpose Interface Bus.  
Types of Messages  
Devices on the GPIB communicate by passing messages through the interface system. There are  
two types of messages:  
Device-dependent messages, often called data or data messages, contain device-specific  
information such as programming instructions, measurement results, machine status, and data  
files.  
Interface messages manage the bus itself. They are usually called commands or command  
messages. Interface messages perform such functions as initializing the bus, addressing and  
unaddressing devices, and setting devices for remote or local programming.  
Note: The term command as used here should not be confused with some device instructions  
which are also referred to as commands. Such device-specific instructions are actually data  
messages.  
Talkers, Listeners, and Controllers  
There are three types of GPIB communicators. A Talker sends data messages to one or more  
Listeners. The Controller manages the flow of information on the GPIB by sending commands to  
all devices.  
Devices can be Talkers, Listeners, and/or Controllers. A digital multimeter, for example, is a  
Talker and may also be a Listener. A printer or plotter is usually only a Listener. A computer on  
the GPIB often combines all three roles to manage the bus and communicate with other devices.  
The GPIB is a bus like a typical computer bus except that the computer has its circuit cards  
interconnected via a backplane bus whereas the GPIB has standalone devices interconnected via a  
cable bus.  
© National Instruments Corporation  
A-1  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation of the GPIB  
Appendix A  
The role of the GPIB Controller can also be compared to the role of the computer's CPU, but a  
better analogy is to the switching center of a city telephone system.  
The switching center (Controller) monitors the communications network (GPIB). When the  
center (Controller) notices that a party (device) wants to make a call (send a data message), it  
connects the caller (Talker) to the receiver (Listener).  
The Controller usually addresses a Talker and a Listener before the Talker can send its message to  
the Listener. After the message is transmitted, the Controller usually unaddresses both devices.  
Some bus configurations do not require a Controller. For example, one device may only be a  
Talker (called a Talk-only device) and there may be one or more Listen-only devices.  
A Controller is necessary when the active or addressed Talker or Listener must be changed. The  
Controller function is usually handled by a computer.  
System Controller and Active Controller  
Although there can be multiple Controllers on the GPIB, only one Controller at a time is Active  
Controller or Controller-in-Charge (CIC). Active control can be passed from the current Active  
Controller to an idle Controller. Only one device on the bus, the System Controller, can make  
itself the Active Controller.  
GPIB Signals  
The interface bus consists of 16 signal lines and 8 ground return or shield drain lines. The 16  
signal lines are divided into three groups:  
8 data lines  
3 handshake lines  
5 interface management lines  
GPIB-100A User Manual  
A-2  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix A  
Operation of the GPIB  
Figure A-1 shows the arrangement of these signals on the GPIB cable connector.  
1
2
3
4
5
6
7
8
9
13  
14  
15  
16  
17  
18  
19  
20  
21  
DIO1*  
DIO2*  
DIO3*  
DIO4*  
EOI*  
DAV*  
NRFD*  
NDAC*  
IFC*  
DIO5*  
DIO6*  
DIO7*  
DIO8*  
REN*  
GND (TW PAIR W/DAV*)  
GND (TW PAIR W/NRFD*)  
GND (TW PAIR W/NDAC*)  
GND (TW PAIR W/IFC*)  
GND (TW PAIR W/SRQ*)  
GND (TW PAIR W/ATN*)  
SIGNAL GROUND  
10 22  
11 23  
12 24  
SRQ*  
ATN*  
SHIELD  
Figure A-1. GPIB Cable Connector  
Data Lines  
The eight data lines, DIO1 through DIO8, carry both data and command messages. All  
commands and most data use the 7-bit ASCII or ISO code set, in which case the eighth bit, DIO8,  
is unused or used for parity.  
Appendix E lists the GPIB command messages.  
Handshake Lines  
Three lines asynchronously control the transfer of message bytes among devices. The process is  
called a three-wire interlocked handshake and it guarantees that message bytes on the data lines are  
sent and received without transmission error.  
NRFD (not ready for data)  
NRFD indicates when a device is ready or not ready to receive a message byte. The line is driven  
by all devices when receiving commands and by Listeners when receiving data messages.  
© National Instruments Corporation  
A-3  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation of the GPIB  
Appendix A  
NDAC (not data accepted)  
NDAC indicates when a device has or has not accepted a message byte. The line is driven by all  
devices when receiving commands and by Listeners when receiving data messages.  
DAV (data valid)  
DAV tells when the signals on the data lines are stable (valid) and can be accepted safely by  
devices. The Controller drives DAV when sending commands, and the Talker drives it when  
sending data messages.  
The way in which NRFD and NDAC are used by the receiving device is called the Acceptor  
Handshake. Likewise, the sending device uses DAV in the Source Handshake.  
Interface Management Lines  
Five lines are used to manage the flow of information across the interface.  
ATN (attention)  
The Controller drives ATN true when it uses the data lines to send commands and false when it  
allows a Talker to send data messages.  
IFC (interface clear)  
The System Controller drives the IFC line to initialize the bus to become Controller-In-Charge.  
REN (remote enable)  
The System Controller drives the REN line, which is used to place devices in remote or local  
program mode.  
SRQ (service request)  
Any device can drive the SRQ line to asynchronously request service from the Active Controller  
with the SRQ line.  
EOI (end or identify)  
The EOI line has two purposes. The Talker uses the EOI line to mark the end of a message string.  
The Active Controller uses the EOI line to tell devices to identify their responses in a parallel poll.  
GPIB-100A User Manual  
A-4  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix A  
Operation of the GPIB  
Physical and Electrical Characteristics  
Devices are usually connected with a cable assembly consisting of a shielded 24-conductor cable  
with both a plug and receptacle at each end. This design allows devices to be connected in either a  
linear or a star configuration, or a combination of the two. See Figures A-2 and A-3.  
Figure A-2. Linear Configuration of the GPIB Devices  
© National Instruments Corporation  
A-5  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation of the GPIB  
Appendix A  
Figure A-3. Star Configuration of GPIB Devices  
The standard connector is the Amphenol or Cinch Series 57 MICRORIBBON or AMP CHAMP  
type. An adapter cable using non-standard cable and/or connector is used for special interconnect  
applications.  
The GPIB uses negative logic with standard TTL logic levels. When DAV is true, for example, it  
is a TTL low level (0.8 V), and when DAV is false, it is a TTL high level (2.0 V).  
GPIB-100A User Manual  
A-6  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix A  
Operation of the GPIB  
Configuration Restrictions  
To achieve the high data transfer rate that the GPIB is designed for, the physical distance between  
devices and the number of devices on the bus is limited.  
The following restrictions are typical:  
A maximum separation of 4 m between any two devices and an average separation of 2 m  
over the entire bus.  
A maximum total cable length of 20 m.  
No more than 15 devices connected to each bus, with at least two-thirds powered-on.  
It is usually possible to connect a cluster of lab instruments without exceeding these restrictions. But  
many applications require longer cable spans or additional loading. From the time the GPIB was  
invented, the need has existed for bus extenders and expanders (repeaters).  
Extenders connect two separate buses via a transmission medium and the distance between the buses  
can be quite long. Expanders allow up to 14 additional devices to be connected to the bus and 20  
meters of cable length to be added to the system.  
National Instruments provides two extenders which allow longer cable spans. These products are the  
GPIB-100A and the GPIB-110. Both must be used in pairs, one at each end of the extension cable.  
The GPIB-100A, a parallel extender, relays the instantaneous status of all GPIB signals over an RS-  
422-compatible cable. The GPIB-100A allows up to a 300-meter extension. The GPIB-110, a serial  
extender, samples the GPIB signals, encodes the information into small packets, and transmits the  
packets on either a low-cost coaxial cable, or a high performance electrically isolated fiber-optic cable.  
The GPIB-110 allows up to a 2-kilometer extension.  
The GPIB-100A is the only parallel extender on the market today. The instantaneous status of all  
GPIB signals on one side are relayed over individual RS-422 circuits to the other side. This  
approach makes the GPIB-100A the fastest and most transparent of all extenders available. The  
parallel design however, requires bulkier and more costly cable than serial designs.  
© National Instruments Corporation  
A-7  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix B  
Schematic Diagram  
This appendix contains the schematic diagram for the GPIB-100A.  
© National Instruments Corporation  
B-1  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix C  
GPIB-100A Parts Locator Diagram  
This appendix contains the parts locator diagram for the GPIB-100A. The parts locator diagram  
shows the locations of the GPIB-100A configuration jumpers and switches.  
Figure C-1. GPIB-100A Parts Locator Diagram  
© National Instruments Corporation  
C-1  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix D  
Cable Assembly Wire List  
This appendix contains the wire list for the GPIB-100A Transmission Cable.  
© National Instruments Corporation  
D-1  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Table D-1. Cable Assembly Wire List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix E  
Multiline Interface Command Messages  
The following tables are multiline interface messages (sent and received with ATN TRUE).  
© National Instruments Corporation  
E-1  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Multiline Interface Command Messages  
Appendix E  
Multiline Interface Messages  
Hex Oct Dec ASCII Msg  
Hex Oct Dec ASCII Msg  
00  
01  
02  
03  
04  
05  
06  
07  
000  
001  
002  
003  
004  
005  
006  
007  
0
1
2
3
4
5
6
7
NUL  
SOH  
STX  
ETX  
EOT  
ENQ  
ACK  
BEL  
20  
21  
22  
23  
24  
25  
26  
27  
040  
041  
042  
043  
044  
045  
046  
047  
32  
33  
34  
35  
36  
37  
38  
39  
SP  
!
"
#
$
%
&
'
MLA0  
MLA1  
MLA2  
MLA3  
MLA4  
MLA5  
MLA6  
MLA7  
GTL  
SDC  
PPC  
08  
09  
010  
011  
012  
013  
014  
015  
016  
017  
8
9
BS  
HT  
LF  
VT  
FF  
CR  
SO  
SI  
GET  
TCT  
28  
29  
050  
051  
052  
053  
054  
055  
056  
057  
40  
41  
42  
43  
44  
45  
46  
47  
(
)
*
+
,
-
.
MLA8  
MLA9  
0A  
0B  
0C  
0D  
0E  
0F  
10  
11  
12  
13  
14  
15  
2A  
2B  
2C  
2D  
2E  
2F  
MLA10  
MLA11  
MLA12  
MLA13  
MLA14  
MLA15  
/
10  
11  
12  
13  
14  
15  
16  
17  
020  
021  
022  
023  
024  
025  
026  
027  
16  
17  
18  
19  
20  
21  
22  
23  
DLE  
DC1  
DC2  
DC3  
DC4  
NAK  
SYN  
ETB  
30  
31  
32  
33  
34  
35  
36  
37  
060  
061  
062  
063  
064  
065  
066  
067  
48  
49  
50  
51  
52  
53  
54  
55  
0
1
2
3
4
5
6
7
MLA16  
MLA17  
MLA18  
MLA19  
MLA20  
MLA21  
MLA22  
MLA23  
LLO  
DCL  
PPU  
18  
19  
030  
031  
032  
033  
034  
035  
036  
037  
24  
25  
26  
27  
28  
29  
30  
31  
CAN  
EM  
SUB  
ESC  
FS  
GS  
RS  
SPE  
SPD  
38  
39  
070  
071  
072  
073  
074  
075  
076  
077  
56  
57  
58  
59  
60  
61  
62  
63  
8
9
:
MLA24  
MLA25  
MLA26  
MLA27  
MLA28  
MLA29  
MLA30  
UNL  
1A  
1B  
1C  
1D  
1E  
1F  
3A  
3B  
3C  
3D  
3E  
3F  
;
<
=
>
?
US  
Message Definitions  
DCL  
GET  
GTL  
LLO  
Device Clear  
Group Execute Trigger  
Go To Local  
MSA My Secondary Address  
MTA My Talk Address  
PPC  
PPD  
Parallel Poll Configure  
Parallel Poll Disable  
Local Lockout  
MLA My Listen Address  
GPIB-100A User Manual  
E-2  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix E  
Multiline Interface Command Messages  
Multiline Interface Messages  
Hex Oct Dec ASCII Msg  
Hex Oct Dec ASCII Msg  
40  
41  
42  
43  
44  
45  
46  
47  
100  
101  
102  
103  
104  
105  
106  
107  
64  
65  
66  
67  
68  
69  
70  
71  
@
A
B
C
D
E
MTA0  
MTA1  
MTA2  
MTA3  
MTA4  
MTA5  
MTA6  
MTA7  
60  
61  
62  
63  
64  
65  
66  
67  
140  
141  
142  
143  
144  
145  
146  
147  
96  
97  
98  
`
MSA0,PPE  
MSA1,PPE  
MSA2,PPE  
MSA3,PPE  
MSA4,PPE  
MSA5,PPE  
MSA6,PPE  
MSA7,PPE  
a
b
c
d
e
f
99  
100  
101  
102  
103  
F
G
g
48  
49  
110  
111  
112  
113  
114  
115  
116  
117  
72  
73  
74  
75  
76  
77  
78  
79  
H
I
J
K
L
M
N
O
MTA8  
MTA9  
68  
69  
150  
151  
152  
153  
154  
155  
156  
157  
104  
105  
106  
107  
108  
109  
110  
111  
h
i
j
k
l
m
n
o
MSA8,PPE  
MSA9,PPE  
MSA10,PPE  
MSA11,PPE  
MSA12,PPE  
MSA13,PPE  
MSA14,PPE  
MSA15,PPE  
4A  
4B  
4C  
4D  
4E  
4F  
MTA10  
MTA11  
MTA12  
MTA13  
MTA14  
MTA15  
6A  
6B  
6C  
6D  
6E  
6F  
50  
51  
52  
53  
54  
55  
56  
57  
120  
121  
122  
123  
124  
125  
126  
127  
80  
81  
82  
83  
84  
85  
86  
87  
P
MTA16  
MTA17  
MTA18  
MTA19  
MTA20  
MTA21  
MTA22  
MTA23  
70  
71  
72  
73  
74  
75  
76  
77  
160  
161  
162  
163  
164  
165  
166  
167  
112  
113  
114  
115  
116  
117  
118  
119  
p
q
r
s
t
u
v
w
MSA16,PPD  
MSA17,PPD  
MSA18,PPD  
MSA19,PPD  
MSA20,PPD  
MSA21,PPD  
MSA22,PPD  
MSA23,PPD  
Q
R
S
T
U
V
W
58  
59  
130  
131  
132  
133  
134  
135  
136  
137  
88  
89  
90  
91  
92  
93  
94  
95  
X
Y
Z
[
\
]
MTA24  
MTA25  
MTA26  
MTA27  
MTA28  
MTA29  
MTA30  
UNT  
78  
79  
170  
171  
172  
173  
174  
175  
176  
177  
120  
121  
122  
123  
124  
125  
126  
127  
x
y
z
{
|
}
MSA24,PPD  
MSA25,PPD  
MSA26,PPD  
MSA27,PPD  
MSA28,PPD  
MSA29,PPD  
MSA30,PPD  
5A  
5B  
5C  
5D  
5E  
5F  
7A  
7B  
7C  
7D  
7E  
7F  
^
_
~
DEL  
PPE  
PPU  
SDC  
SPD  
Parallel Poll Enable  
Parallel Poll Unconfigure  
Selected Device Clear  
Serial Poll Disable  
SPE  
Serial Poll Enable  
Take Control  
Unlisten  
TCT  
UNL  
UNT  
Untalk  
© National Instruments Corporation  
E-3  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix F  
Mnemonics Key  
This appendix contains a mnemonics key that defines the mnemonics (abbreviations) used  
throughout this manual.  
Mnemonic  
Definition  
ASCII  
American Standard Code for Information Interchange  
ATN  
C
Attention  
Controller  
CIC  
DAV  
DIO  
Controller-In-Charge Bit  
Data Valid  
Data  
EOI  
FF  
End or Identify Bit  
Flip-flop  
IDY  
IFC  
Identify  
Interface Clear  
ISO  
L
International Standard code set  
Listener  
LAC  
LS  
Local Active Controller  
Local Source  
LSC  
NDAC  
NRFD  
PON  
PP  
Local System Controller  
GPIB Not Data Accepted line status Bit  
GPIB Not Ready For Data line status Bit  
Power-On Reset Pulse  
Parallel Poll (scan all status flags)  
Parallel Poll Response  
Remote Active Controller  
Resistor/Capacitor Delay  
PPR  
RAC  
RCD  
© National Instruments Corporation  
F-1  
GPIB-100A User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Mnemonics Key  
Appendix F  
Mnemonic  
REN  
RFD  
RR  
Definition  
Remote Enable  
Ready for Data  
Remote Response  
Remote Source  
RS  
RSC  
SRQ  
T
Remote System Controller  
Service Request  
Talker  
TL  
Talker/Listener  
TLC  
TTL  
Talker/Listener/Controller (GPIB Adapter)  
Transistor/Transistor Logic  
GPIB-100A User Manual  
F-2  
© National Instruments Corporation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
User Comment Form  
National Instruments encourages you to comment on the documentation supplied with our  
products. This information helps us provide quality products to meet your needs.  
Title: GPIB-100A User Manual  
Edition Date  
Part Number:  
March 1990  
320063-01  
Please comment on the completeness, clarity, and organization of the manual.  
If you find errors in the manual, please record the page numbers and describe the errors.  
Thank you for your help.  
Name  
Title  
Company  
Address  
Phone  
(
)
Mail to: Technical Publications  
National Instruments Corporation  
6504 Bridge Point Parkway, MS 53-02  
Austin, TX 78730-5039  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Miele Coffeemaker CVA 2662 User Manual
Miele Refrigerator KF 1801 SF User Manual
Minolta Universal Remote SRU 3040 10 User Manual
Omron Healthcare Blood Pressure Monitor BP742 User Manual
Oregon Scientific Blood Pressure Monitor BPW810 User Manual
Outback Power Systems Power Supply FX2012MT User Manual
Palsonic Flat Panel Television TFTV3839DT User Manual
Palsonic TV Converter Box 3425G User Manual
Panasonic Fax Machine KX F900 User Manual
Panasonic Photo Printer AL TA10U User Manual