National Instruments Network Card DAQ S User Manual

DAQ S Series  
NI 6124/6154 User Manual  
DAQ-STC2 S Series Simultaneous Sampling Multifunction Input/Output Devices  
NI 6124/6154 User Manual  
August 2008  
372613A-01  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Important Information  
Warranty  
NI 6124 and NI 6154 devices are warranted against defects in materials and workmanship for a period of one year from the date of shipment,  
as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective  
during the warranty period. This warranty includes parts and labor.  
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects in  
materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National Instruments  
will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives notice of such defects  
during the warranty period. National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.  
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any  
equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by  
warranty.  
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical accuracy. In  
the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document  
without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected. In no event shall National  
Instruments be liable for any damages arising out of or related to this document or the information contained in it.  
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF  
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMERS RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF NATIONAL  
INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING  
FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY THEREOF. This limitation of  
the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including negligence. Any action against  
National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in  
performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service  
failures caused by owner’s failure to follow the National Instruments installation, operation, or maintenance instructions; owner’s modification of the  
product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside  
reasonable control.  
Copyright  
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying,  
recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National  
Instruments Corporation.  
National Instruments respects the intellectual property of others, and we ask our users to do the same. NI software is protected by copyright and other  
intellectual property laws. Where NI software may be used to reproduce software or other materials belonging to others, you may use NI software only  
to reproduce materials that you may reproduce in accordance with the terms of any applicable license or other legal restriction.  
Trademarks  
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section  
on ni.com/legalfor more information about National Instruments trademarks.  
Other product and company names mentioned herein are trademarks or trade names of their respective companies.  
Members of the National Instruments Alliance Partner Program are business entities independent from National Instruments and have no agency,  
partnership, or joint-venture relationship with National Instruments.  
Patents  
For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software,  
the patents.txtfile on your media, or the National Instruments Patent Notice at ni.com/patents.  
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS  
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF  
RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN  
ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT  
INJURY TO A HUMAN.  
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE  
IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY,  
COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS  
AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND HARDWARE  
COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL DEVICES,  
TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR  
ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE HEREAFTER  
COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD CREATE A RISK OF  
HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD NOT BE RELIANT SOLELY  
UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID DAMAGE, INJURY, OR DEATH,  
THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES,  
INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS  
CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING PLATFORMS AND BECAUSE A USER OR APPLICATION  
DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT  
EVALUATED OR CONTEMPLATED BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY  
RESPONSIBLE FOR VERIFYING AND VALIDATING THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER  
NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT  
LIMITATION, THE APPROPRIATE DESIGN, PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Compliance  
Compliance with FCC/Canada Radio Frequency Interference  
Regulations  
Determining FCC Class  
The Federal Communications Commission (FCC) has rules to protect wireless communications from interference. The FCC  
places digital electronics into two classes. These classes are known as Class A (for use in industrial-commercial locations only)  
or Class B (for use in residential or commercial locations). All National Instruments (NI) products are FCC Class A products.  
Depending on where it is operated, this Class A product could be subject to restrictions in the FCC rules. (In Canada, the  
Department of Communications (DOC), of Industry Canada, regulates wireless interference in much the same way.) Digital  
electronics emit weak signals during normal operation that can affect radio, television, or other wireless products.  
All Class A products display a simple warning statement of one paragraph in length regarding interference and undesired  
operation. The FCC rules have restrictions regarding the locations where FCC Class A products can be operated.  
Consult the FCC Web site at www.fcc.govfor more information.  
FCC/DOC Warnings  
This equipment generates and uses radio frequency energy and, if not installed and used in strict accordance with the instructions  
in this manual and the CE marking Declaration of Conformity*, may cause interference to radio and television reception.  
Classification requirements are the same for the Federal Communications Commission (FCC) and the Canadian Department  
of Communications (DOC).  
Changes or modifications not expressly approved by NI could void the user’s authority to operate the equipment under the  
FCC Rules.  
Class A  
Federal Communications Commission  
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC  
Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated  
in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and  
used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this  
equipment in a residential area is likely to cause harmful interference in which case the user is required to correct the interference  
at their own expense.  
Canadian Department of Communications  
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.  
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.  
Compliance with EU Directives  
Users in the European Union (EU) should refer to the Declaration of Conformity (DoC) for information* pertaining to the  
CE marking. Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory compliance  
information. To obtain the DoC for this product, visit ni.com/certification, search by model number or product line,  
and click the appropriate link in the Certification column.  
*
The CE marking Declaration of Conformity contains important supplementary information and instructions for the user or  
installer.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
About This Manual  
Chapter 1  
Installing NI-DAQmx....................................................................................................1-1  
Installing Other Software...............................................................................................1-1  
Device Self-Calibration .................................................................................................1-2  
Chapter 2  
Calibration Circuitry......................................................................................................2-4  
Internal or Self-Calibration..............................................................................2-4  
Signal Conditioning .......................................................................................................2-5  
Chapter 3  
NI 6124 I/O Connector Signal Descriptions..................................................................3-1  
Chapter 4  
Analog Input Terminal Configuration ...........................................................................4-2  
Input Polarity and Range ...............................................................................................4-3  
Working Voltage Range ................................................................................................4-4  
AI Data Acquisition Methods........................................................................................4-4  
Analog Input Triggering ................................................................................................4-6  
© National Instruments Corporation  
v
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Routing AI Start Trigger to an Output Terminal.............................. 4-19  
AI Reference Trigger Signal........................................................................... 4-19  
Using an Analog Source................................................................... 4-20  
Chapter 5  
Connecting Analog Output Signals............................................................................... 5-4  
Waveform Generation Timing Signals.......................................................................... 5-6  
AO Sample Clock Signal ................................................................................ 5-6  
Using an Internal Source .................................................................. 5-6  
Using an External Source ................................................................. 5-7  
Routing AO Sample Clock Signal to an Output Terminal ............... 5-7  
NI 6124/6154 User Manual  
vi  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using an Analog Source ...................................................................5-10  
Routing AO Start Trigger Signal to an Output Terminal..................5-10  
Using a Digital Source ......................................................................5-11  
Chapter 6  
Getting Started with DIO Applications in Software on Non-Isolated Devices.....6-10  
Digital I/O for Isolated Devices.....................................................................................6-11  
I/O Protection for Isolated Devices.................................................................6-12  
Chapter 7  
Counters  
Single Point (On-Demand) Edge Counting ......................................7-2  
Buffered (Sample Clock) Edge Counting .........................................7-3  
Controlling the Direction of Counting..............................................7-4  
Pulse-Width Measurement ..............................................................................7-4  
Single Pulse-Width Measurement.....................................................7-4  
Buffered Pulse-Width Measurement.................................................7-5  
© National Instruments Corporation  
vii  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Routing Counter n Z Signal to an Output Terminal ......................... 7-28  
Counter n Up_Down Signal............................................................................ 7-29  
Counter n HW Arm Signal.............................................................................. 7-29  
Routing Signals to Counter n HW Arm Input .................................. 7-29  
Counter n Internal Output and Counter n TC Signals..................................... 7-29  
Routing Counter n Internal Output to an Output Terminal .............. 7-30  
NI 6124/6154 User Manual  
viii  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
When To Use Duplicate Count Prevention.......................................7-37  
Enabling Duplicate Count Prevention in NI-DAQmx ......................7-37  
80 MHz Source Mode.......................................................................7-38  
Chapter 8  
Exporting Timing Output Signals Using PFI Terminals ...............................................8-4  
Using PFI Terminals as Static Digital Inputs and Outputs............................................8-5  
PFI Filters ......................................................................................................................8-6  
Chapter 9  
80 MHz Timebase ...........................................................................................9-2  
20 MHz Timebase ...........................................................................................9-2  
100 kHz Timebase...........................................................................................9-2  
External Reference Clock................................................................................9-2  
10 MHz Reference Clock................................................................................9-3  
Synchronizing Multiple Devices ...................................................................................9-3  
© National Instruments Corporation  
ix  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
PXI Clock and Trigger Signals...................................................................................... 9-8  
PXI_CLK10 .................................................................................................... 9-8  
PXI_STAR Trigger......................................................................................... 9-8  
MITE and DAQ-PnP..................................................................................................... 10-1  
PXI Considerations........................................................................................................ 10-1  
PXI Express..................................................................................................... 10-1  
Triggering with a Digital Source................................................................................... 11-1  
Triggering with an Analog Source ................................................................................ 11-2  
Analog Trigger Actions .................................................................................. 11-3  
Analog Trigger Types.................................................................................................... 11-3  
Analog Trigger Accuracy.............................................................................................. 11-6  
Appendix A  
Technical Support and Professional Services  
Glossary  
Index  
NI 6124/6154 User Manual  
x
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
About This Manual  
The NI 6124/6154 User Manual contains information about using the  
National Instruments S Series NI 6124 and NI 6154 data acquisition (DAQ)  
devices with NI-DAQmx 8.8 and later.  
Conventions  
The following conventions appear in this manual:  
<>  
Angle brackets that contain numbers separated by an ellipsis represent  
a range of values associated with a bit or signal name—for example,  
AO <3..0>.  
»
The » symbol leads you through nested menu items and dialog box options  
to a final action. The sequence File»Page Setup»Options directs you to  
pull down the File menu, select the Page Setup item, and select Options  
from the last dialog box.  
This icon denotes a note, which alerts you to important information.  
This icon denotes a caution, which advises you of precautions to take to  
avoid injury, data loss, or a system crash. When this symbol is marked on a  
product, refer to the Read Me First: Safety and Radio-Frequency  
Interference document for information about precautions to take.  
bold  
Bold text denotes items that you must select or click in the software, such  
as menu items and dialog box options. Bold text also denotes parameter  
names.  
italic  
Italic text denotes variables, emphasis, a cross-reference, or an introduction  
to a key concept. Italic text also denotes text that is a placeholder for a word  
or value that you must supply.  
monospace  
Text in this font denotes text or characters that you should enter from the  
keyboard, sections of code, programming examples, and syntax examples.  
This font is also used for the proper names of disk drives, paths, directories,  
programs, subprograms, subroutines, device names, functions, operations,  
variables, filenames, and extensions.  
Platform  
Text in this font denotes a specific platform and indicates that the text  
following it applies only to that platform.  
© National Instruments Corporation  
xi  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
About This Manual  
Related Documentation  
Each application software package and driver includes information about  
writing applications for taking measurements and controlling measurement  
devices. The following references to documents assume you have  
NI-DAQmx 8.8 or later, and where applicable, version 7.1 or later of the  
NI application software.  
NI-DAQmx for Windows  
The DAQ Getting Started Guide describes how to install your NI-DAQmx  
for Windows software, how to install your NI-DAQmx-supported DAQ  
device, and how to confirm that your device is operating properly. Select  
Start»All Programs»National Instruments»NI-DAQ»DAQ Getting  
Started Guide.  
The NI-DAQ Readme lists which devices are supported by this version of  
NI-DAQmx. Select Start»All Programs»National Instruments»  
NI-DAQ»NI-DAQ Readme.  
The NI-DAQmx Help contains general information about measurement  
concepts, key NI-DAQmx concepts, and common applications that are  
applicable to all programming environments. Select Start»All Programs»  
National Instruments»NI-DAQ»NI-DAQmx Help.  
LabVIEW  
If you are a new user, use the Getting Started with LabVIEW manual to  
familiarize yourself with the LabVIEW graphical programming  
environment and the basic LabVIEW features you use to build data  
acquisition and instrument control applications. Open the Getting Started  
with LabVIEW manual by selecting Start»All Programs»National  
Instruments»LabVIEW»LabVIEW Manuals or by navigating to the  
labview\manualsdirectory and opening LV_Getting_Started.pdf.  
Use the LabVIEW Help, available by selecting Help»Search the  
LabVIEW Help in LabVIEW, to access information about LabVIEW  
programming concepts, step-by-step instructions for using LabVIEW, and  
reference information about LabVIEW VIs, functions, palettes, menus, and  
tools. Refer to the following locations on the Contents tab of the LabVIEW  
Help for information about NI-DAQmx:  
Getting Started with LabVIEW»Getting Started with DAQ—  
Includes overview information and a tutorial to learn how to take an  
NI-DAQmx measurement in LabVIEW using the DAQ Assistant.  
NI 6124/6154 User Manual  
xii  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
About This Manual  
VI and Function Reference»Measurement I/O VIs and  
Functions—Describes the LabVIEW NI-DAQmx VIs and properties.  
Taking Measurements—Contains the conceptual and how-to  
information you need to acquire and analyze measurement data in  
LabVIEW, including common measurements, measurement  
fundamentals, NI-DAQmx key concepts, and device considerations.  
LabWindows/CVI  
The Data Acquisition book of the LabWindows/CVI Help contains  
measurement concepts for NI-DAQmx. This book also contains Taking an  
NI-DAQmx Measurement in LabWindows/CVI, which includes  
step-by-step instructions about creating a measurement task using the DAQ  
Assistant. In LabWindows/CVI, select Help»Contents, then select  
Using LabWindows/CVI»Data Acquisition.  
The NI-DAQmx Library book of the LabWindows/CVI Help contains  
API overviews and function reference for NI-DAQmx. Select Library  
Reference»NI-DAQmx Library in the LabWindows/CVI Help.  
Measurement Studio  
If you program your NI-DAQmx-supported device in Measurement Studio  
using Visual C++, Visual C#, or Visual Basic .NET, you can interactively  
create channels and tasks by launching the DAQ Assistant from MAX or  
from within Visual Studio .NET. You can generate the configuration code  
based on your task or channel in Measurement Studio. Refer to the DAQ  
Assistant Help for additional information about generating code. You also  
can create channels and tasks, and write your own applications in your  
ADE using the NI-DAQmx API.  
For help with NI-DAQmx methods and properties, refer to the NI-DAQmx  
.NET Class Library or the NI-DAQmx Visual C++ Class Library included  
in the NI Measurement Studio Help. For general help with programming in  
Measurement Studio, refer to the NI Measurement Studio Help, which is  
fully integrated with the Microsoft Visual Studio .NET help. To view  
this help file in Visual Studio. NET, select Measurement Studio»  
NI Measurement Studio Help.  
© National Instruments Corporation  
xiii  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
About This Manual  
To create an application in Visual C++, Visual C#, or Visual Basic .NET,  
follow these general steps:  
1. In Visual Studio .NET, select File»New»Project to launch the New  
Project dialog box.  
2. Find the Measurement Studio folder for the language you want to  
create a program in.  
3. Choose a project type. You add DAQ tasks as a part of this step.  
ANSI C without NI Application Software  
The NI-DAQmx Help contains API overviews and general information  
about measurement concepts. Select Start»All Programs»National  
Instruments»NI-DAQ»NI-DAQmx Help.  
The NI-DAQmx C Reference Help describes the NI-DAQmx Library  
functions, which you can use with National Instruments data acquisition  
devices to develop instrumentation, acquisition, and control applications.  
Select Start»All Programs»National Instruments»NI-DAQ»  
NI-DAQmx C Reference Help.  
.NET Languages without NI Application Software  
With the Microsoft .NET Framework version 1.1 or later, you can use  
NI-DAQmx to create applications using Visual C# and Visual Basic .NET  
without Measurement Studio. You need Microsoft Visual Studio  
.NET 2003 or Microsoft Visual Studio 2005 for the API documentation to  
be installed.  
The installed documentation contains the NI-DAQmx API overview,  
measurement tasks and concepts, and function reference. This help is fully  
integrated into the Visual Studio .NET documentation. To view the  
NI-DAQmx .NET documentation, go to Start»Programs»National  
Instruments»NI-DAQ»NI-DAQmx .NET Reference Help. Expand  
NI Measurement Studio Help»NI Measurement Studio .NET Class  
Library»Reference to view the function reference. Expand  
NI Measurement Studio Help»NI Measurement Studio .NET Class  
Library»Using the Measurement Studio .NET Class Libraries to view  
conceptual topics for using NI-DAQmx with Visual C# and Visual Basic  
.NET.  
To get to the same help topics from within Visual Studio, go to Help»  
Contents. Select Measurement Studio from the Filtered By drop-down  
list and follow the previous instructions.  
NI 6124/6154 User Manual  
xiv  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
About This Manual  
Device Documentation and Specifications  
The NI 6124 Specifications and NI 6154 Specifications documents contain  
all specifications for the NI 6124 and NI 6154 S Series devices respectively.  
Documentation for supported devices and accessories, including PDF  
and help files describing device terminals, specifications, features, and  
operation are on the NI-DAQmx CD that includes Device Documentation.  
Insert the CD, open the Device Documentation directory, and double-click  
the Device Documents shortcut for your language to find, view, and print  
device documents.  
Training Courses  
If you need more help getting started developing an application with  
NI products, NI offers training courses. To enroll in a course or obtain a  
detailed course outline, refer to ni.com/training.  
Technical Support on the Web  
For additional support, refer to ni.com/supportor zone.ni.com.  
Note You can download these documents at ni.com/manuals.  
DAQ specifications and some DAQ manuals are available as PDFs. You  
must have Adobe Acrobat Reader with Search and Accessibility 5.0.5 or  
later installed to view the PDFs. Refer to the Adobe Systems Incorporated  
Web site at www.adobe.comto download Acrobat Reader. Refer to the  
National Instruments Product Manuals Library at ni.com/manualsfor  
updated documentation resources.  
© National Instruments Corporation  
xv  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
1
Getting Started  
The NI 6124 and NI 6154 are simultaneous sampling multifunction I/O  
devices (S Series) that use the DAQ-STC2 ASIC.  
The NI 6124 S Series is a non-isolated device featuring PXI Express  
connectivity, four simultaneously sampling 16-bit analog inputs,  
two 16-bit voltage analog outputs, 24 lines of bidirectional DIO,  
and two general-purpose 32-bit counter/timers.  
The NI 6154 S Series is an isolated PCI device featuring four isolated  
differential 16-bit analog inputs, four isolated 16-bit analog outputs,  
six DI lines, four DO lines, and two general-purpose 32-bit counter/timers.  
If you have not already installed your device, refer to the DAQ Getting  
Started Guide. For specifications arranged by S Series device family, refer  
to the specifications document for your device on ni.com/manuals.  
Before installing your DAQ device, you must install the software you plan  
to use with the device.  
Installing NI-DAQmx  
The DAQ Getting Started Guide, which you can download at ni.com/  
manuals, offers NI-DAQmx users step-by-step instructions for installing  
software and hardware, configuring channels and tasks, and getting started  
developing an application.  
Installing Other Software  
If you are using other software, refer to the installation instructions that  
accompany your software.  
© National Instruments Corporation  
1-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 1  
Getting Started  
Installing the Hardware  
The DAQ Getting Started Guide contains non-software-specific  
information about how to install PCI and PXI Express devices, as well as  
accessories and cables.  
Device Self-Calibration  
NI recommends that you self-calibrate your S Series device after  
installation and whenever the ambient temperature changes.  
Self-calibration should be performed after the device has warmed up for the  
recommended time period. Refer to the device specifications to find your  
device warm-up time. This function measures the onboard reference  
voltage of the device and adjusts the self-calibration constants to account  
for any errors caused by short-term fluctuations in the environment.  
Disconnect all external signals when you self-calibrate a device.  
You can initiate self-calibration using Measurement & Automation  
Explorer (MAX), by completing the following steps.  
1. Launch MAX.  
2. Select My System»Devices and Interfaces»NI-DAQmx  
Devices»your device.  
3. Initiate self-calibration using one of the following methods:  
Click Self-Calibrate in the upper right corner of MAX.  
Right-click the name of the device in the MAX configuration tree  
and select Self-Calibrate from the drop-down menu.  
Note You can also programmatically self-calibrate your device with NI-DAQmx,  
as described in Device Calibration in the NI-DAQmx Help or the LabVIEW Help in  
version 8.0 or later.  
NI 6124/6154 User Manual  
1-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 1  
Getting Started  
Device Pinouts  
Refer to Appendix A, Device-Specific Information, for NI 6124 and  
NI 6154 device pinouts.  
Device Specifications  
Refer to the specifications for your device, the NI 6124 Specifications or  
the NI 6154 Specifications, available on the NI-DAQ Device Document  
Browser or ni.com/manuals, for more detailed information about the  
NI 6124 and NI 6154 devices.  
© National Instruments Corporation  
1-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
2
DAQ System Overview  
Figure 2-1 shows a typical DAQ system setup, which includes transducers,  
signal conditioning, cables that connect the various devices to the  
accessories, the S Series device, and the programming software. Refer to  
Appendix A, Device-Specific Information, for a list of devices and their  
compatible accessories.  
4
3
2
+
V
1
5
1
2
3
Sensors and Transducers  
Signal Conditioning  
Cable Assembly  
4
5
DAQ Hardware  
Personal Computer/Chassis  
and DAQ Software  
Figure 2-1. Typical DAQ System  
© National Instruments Corporation  
2-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 2  
DAQ System Overview  
DAQ Hardware  
DAQ hardware digitizes signals, performs D/A conversions to generate  
analog output signals, and measures and controls digital I/O signals. The  
following sections contain more information about specific components of  
the DAQ hardware.  
Figure 2-2 shows the components of the non-isolated S Series (NI 6124)  
device.  
Analog Input  
Analog Output  
Digital I/O  
Counters  
PFI  
Bus  
Interface  
Digital  
Routing  
Bus  
RTSI  
Figure 2-2. General NI 6124 Block Diagram  
NI 6124/6154 User Manual  
2-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 2  
DAQ System Overview  
(NI 6154 Only) S Series isolated hardware also includes bank and  
channel-to-channel isolation. Isolated DAQ hardware allows for increased  
protection against hazardous voltages, rejection of common-mode  
voltages, and reduction of ground loops and their associated noise.  
Figure 2-3 shows the components of the isolated S Series (NI 6154) device.  
A
Isolation  
Barrier  
Analog Input  
A
AI GND  
Analog Output  
Counters  
A
AO GND  
Digital  
Isolators  
Digital  
Routing  
Bus  
Interface  
Bus  
PFI/Static DI  
PFI/Static DO  
RTSI  
P0  
P1  
P0.GND  
P1.GND  
Figure 2-3. General NI 6154 Block Diagram  
DAQ-STC2  
The DAQ-STC2 implements a high-performance digital engine for  
S Series data acquisition hardware. Some key features of this engine  
include the following:  
Flexible AI and AO sample and convert timing  
Many triggering modes  
Independent AI, AO, DI, and DO FIFOs  
Generation and routing of RTSI signals for multi-device  
synchronization  
Generation and routing of internal and external timing signals  
© National Instruments Corporation  
2-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 2  
DAQ System Overview  
Two flexible 32-bit counter/timer modules with hardware gating  
Digital waveform acquisition and generation  
Static DIO signals  
True 5 V high current drive DO  
PLL for clock synchronization  
PCI/PXI interface  
Independent scatter-gather DMA controllers for all acquisition and  
generation functions  
Calibration Circuitry  
Calibration is the process of making adjustments to a measurement device  
to reduce errors associated with measurements. Without calibration, the  
measurement results of your device will drift over time and temperature.  
Calibration adjusts for these changes to improve measurement accuracy  
and ensure that your product meets its required specifications.  
DAQ devices have high precision analog circuits that must be adjusted to  
obtain optimum accuracy in your measurements. Calibration determines  
what adjustments these analog circuits should make to the device  
measurements. During calibration, the value of a known, high precision  
measurement source is compared to the value your device acquires or  
generates. The adjustment values needed to minimize the difference  
between the known and measured values are stored in the EEPROM of the  
device as calibration constants. Before performing a measurement, these  
constants are read out of the EEPROM and are used to adjust the calibration  
hardware on the device. NI-DAQmx determines when this is necessary and  
does it automatically. If you are not using NI-DAQmx, you must load these  
values yourself.  
You can calibrate S Series devices in two ways—through internal  
(or self-calibration) or through external calibration.  
Internal or Self-Calibration  
Self-calibration is a process to adjust the device relative to a highly accurate  
and stable internal reference on the device. Self-calibration is similar to the  
autocalibration or autozero found on some instruments. You should  
perform a self-calibration whenever environmental conditions, such as  
ambient temperature, change significantly. To perform self-calibration, use  
the self-calibrate function or VI that is included with your driver software.  
Self-calibration requires no external connections.  
NI 6124/6154 User Manual  
2-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 2  
DAQ System Overview  
External Calibration  
External calibration is a process to adjust the device relative to a traceable,  
high precision calibration standard. The accuracy specifications of your  
device change depending on how long it has been since your last external  
calibration. National Instruments recommends that you calibrate your  
device at least as often as the intervals listed in the accuracy specifications.  
For a detailed calibration procedure for NI 6154 S Series devices, refer to  
the Isolated M/S Series Calibration Procedure, which you can find at  
ni.com/calibrationand selecting Manual Calibration Procedures.  
Signal Conditioning  
Many sensors and transducers require signal conditioning before a  
computer-based measurement system can effectively and accurately  
acquire the signal. The front-end signal conditioning system can include  
functions such as signal amplification, attenuation, filtering, electrical  
isolation, simultaneous sampling, and multiplexing. In addition, many  
transducers require excitation currents or voltages, bridge completion,  
linearization, or high amplification for proper and accurate operation.  
Therefore, most computer-based measurement systems include some form  
of signal conditioning in addition to plug-in data acquisition DAQ devices.  
Sensors and Transducers  
Sensors can generate electrical signals to measure physical phenomena,  
such as temperature, force, sound, or light. Some commonly used sensors  
are strain gages, thermocouples, thermistors, angular encoders, linear  
encoders, and resistance temperature detectors (RTDs).  
To measure signals from these various transducers, you must convert them  
into a form that a DAQ device can accept. For example, the output voltage  
of most thermocouples is very small and susceptible to noise. Therefore,  
you may need to amplify or filter the thermocouple output before digitizing  
it. The manipulation of signals to prepare them for digitizing is called  
signal conditioning.  
For more information about sensors, refer to the following documents.  
For general information about sensors, visit ni.com/sensors.  
If you are using LabVIEW, refer to the LabVIEW Help by selecting  
Help»Search the LabVIEW Help in LabVIEW and then navigate to  
the Taking Measurements book on the Contents tab.  
© National Instruments Corporation  
2-5  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 2  
DAQ System Overview  
If you are using other application software, refer to Common Sensors  
in the NI-DAQmx Help or the LabVIEW Help in version 8.0 or later.  
Programming Devices in Software  
National Instruments measurement devices are packaged with NI-DAQmx  
driver software, an extensive library of functions and VIs you can call from  
your application software, such as LabVIEW or LabWindows/CVI, to  
program all the features of your NI measurement devices. NI-DAQmx  
driver software has an application programming interface (API), which is  
a library of VIs, functions, classes, attributes, and properties for creating  
applications for your device.  
NI-DAQ includes two NI-DAQ drivers—Traditional NI-DAQ (Legacy) and  
NI-DAQmx. DAQ-STC2-based S Series devices use the NI-DAQmx driver.  
Each driver has its own API, hardware configuration, and software  
configuration. Refer to the DAQ Getting Started Guide for more  
information about the two drivers.  
NI-DAQmx includes a collection of programming examples to help you get  
started developing an application. You can modify example code and save  
it in an application. You can use examples to develop a new application or  
add example code to an existing application.  
To locate LabVIEW and LabWindows/CVI examples, open the National  
Instruments Example Finder. In LabVIEW and LabWindows/CVI, select  
Help»Find Examples.  
Measurement Studio, Visual Basic, and ANSI C examples are in the  
following directories:  
NI-DAQmx examples for Measurement Studio-supported languages  
are in the following directories:  
– MeasurementStudio\VCNET\Examples\NIDaq  
– MeasurementStudio\DotNET\Examples\NIDaq  
NI-DAQmx examples for ANSI C are in the NI-DAQ\Examples\  
DAQmx ANSI C Devdirectory  
For additional examples, refer to zone.ni.com.  
NI 6124/6154 User Manual  
2-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
3
I/O Connector  
This chapter contains information about the S Series I/O connector. Refer  
to one of the following sections, depending on your device:  
NI 6124 I/O Connector Signal Descriptions  
NI 6154 I/O Connector Signal Descriptions  
Refer to Appendix A, Device-Specific Information, for the I/O connector  
pinout for your device.  
NI 6124 I/O Connector Signal Descriptions  
(NI 6124 Only) Table 3-1 describes the signals found on the NI 6124 I/O  
connector. For more information about these signals, refer to the NI 6124  
Specifications.  
Table 3-1. NI 6124 Device Signal Descriptions  
I/O Connector Pin  
Reference  
Direction  
Signal Description  
AI <0..3> GND  
Analog Input Channels 0 through 3 Ground—These  
pins are the bias current return point for differential  
measurements.  
AI <0..3> +  
AI <0..3> –  
AO <0..1>  
AI <0..3> GND  
AI <0..3> GND  
AO GND  
Input  
Input  
Analog Input Channels 0 through 3 (+)—These pins  
are routed to the (+) terminal of the respective channel  
amplifier.  
Analog Input Channels 0 through 3 (–)—These pins  
are routed to the (–) terminal of the respective channel  
amplifier.  
Output  
Analog Output Channels 0 through 1—These pins  
supply the voltage output of analog output channels 0  
and 1.  
AO GND  
D GND  
Analog Output Ground—The AO voltages and the  
external reference voltage are referenced to these pins.  
Digital Ground—These pins supply the reference for the  
digital signals at the I/O connector and the +5 VDC  
supply.  
© National Instruments Corporation  
3-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
               
Chapter 3  
I/O Connector  
Table 3-1. NI 6124 Device Signal Descriptions (Continued)  
I/O Connector Pin  
P0.<0..7>  
Reference  
D GND  
Direction  
Signal Description  
Input or Output Digital I/O Channels 0 through 7You can  
individually configure each signal as an input or output.  
P0.6 and P0.7 can also control the up/down signal of  
Counters 0 and 1, respectively.  
PFI <0..7>/P1.<0..7>  
PFI <8..15>/P2.<0..7>  
D GND  
Input or Output Programmable Function Interface or Digital I/O  
Channels 0 through 7 and Channels 8 through  
15—Each of these terminals can be individually  
configured as a PFI terminal or a digital I/O terminal.  
As an input, each PFI terminal can be used to supply an  
external source for AI, AO, DI, and DO timing signals or  
counter/timer inputs.  
As a PFI output, you can route many different internal AI,  
AO, DI, or DO timing signals to each PFI terminal. You  
also can route the counter/timer outputs to each PFI  
terminal.  
As a Port 1 or Port 2 digital I/O signal, you can  
individually configure each signal as an input or output.  
+5 V  
D GND  
Output  
+5 Power Source—These pins provide +5 V power. For  
more information, refer to the +5 V Power Source  
section.  
NI 6154 I/O Connector Signal Descriptions  
(NI 6154 Only) Table 3-2 describes the signals found on the NI 6154 I/O  
connector. For more information about these signals, refer to the NI 6154  
Specifications.  
Table 3-2. NI 6154 I/O Connector Signal Descriptions  
I/O Connector Pin  
Reference  
Direction  
Signal Description  
AI <0..3> +  
AI <0..3> –  
Input  
Analog Input Channels 0 through 3 (+)—These pins are  
routed to the (+) terminal of the respective channel  
amplifier.  
AI <0..3> –  
AO <0..3> +  
Input  
Analog Input Channels 0 through 3 (–)—The reference  
pins for the corresponding AI <0..3> + pin.  
AO <0..3> –  
Output  
Analog Output Channels 0 through 3 (+)—These pins  
supply the voltage output of Analog Output channels 0  
through 3.  
AO <0..3> –  
Output  
Analog Output Channels 0 through 3 (–)—The  
reference pins for the corresponding AO <0..3> + pin.  
NI 6124/6154 User Manual  
3-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 3  
I/O Connector  
Table 3-2. NI 6154 I/O Connector Signal Descriptions (Continued)  
I/O Connector Pin  
Reference  
Direction  
Signal Description  
PFI <0..5>/P0.<0..5> D GND  
Input  
Programmable Function Interface or Static Digital  
Input Channels 0 to 5—Each of these terminals can be  
individually configured as a PFI terminal or a digital input  
terminal.  
As an input, each PFI terminal can be used to supply an  
external source for AI or AO timing signals or  
counter/timer inputs.  
Note: PFI <0..5>/P0.<0..5> are isolated from earth  
ground and chassis ground.  
PFI <6..9>/P1.<0..3> D GND  
Output  
Programmable Function Interface or Static Digital  
Output Channels 6 to 9—Each of these terminals can be  
individually configured as a PFI terminal or a digital  
output terminal.  
As a PFI output, you can route many different internal AI  
or AO timing signals to each PFI terminal. You also can  
route the counter/timer outputs to each PFI terminal.  
Note: PFI <6..9>/P1.<0..3> are isolated from earth  
ground and chassis ground.  
D GND  
Digital Ground—D GND supplies the reference  
for input PFI <0..5>/P0.<0..5> and output PFI <6..9>/  
P1.<0..3>.  
Note: D GND is isolated from earth ground and chassis  
ground.  
NC  
No Connect—Do not connect signals to these terminals.  
+5 V Power Source  
(NI 6124 Only) The +5 V pins on the I/O connector supply +5 V power. You  
can use these pins, referenced to D GND, to power external circuitry.  
Power rating (most devices): +4.65 to +5.25 VDC at 1 A.  
To find your device’s power rating, refer to the specifications document for  
your device.  
Caution Never connect these +5 V power pins to analog or digital ground or to any other  
voltage source on the S Series device or any other device. Doing so can damage the device  
and the computer. NI is not liable for damage resulting from such a connection.  
© National Instruments Corporation  
3-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
4
Analog Input  
Figure 4-1 shows the analog input circuitry of each channel of the  
non-isolated S Series (NI 6124) device.  
Instrumentation  
Amplifier  
Filter  
AI+  
AI–  
AI FIFO  
AI Data  
ADC  
Mux  
Analog  
Trigger  
CAL  
AI Timing Signals  
Figure 4-1. Non-Isolated S Series Analog Input Block Diagram  
Figure 4-2 shows the analog input circuitry of each channel of the isolated  
S Series (NI 6154) device.  
Isolation  
Barrier  
Instrumentation  
Amplifier  
Filter  
AI+  
AI–  
ADC  
AI Data  
AI FIFO  
Digital  
Isolators  
Mux  
AI Timing Signals  
CAL  
Figure 4-2. Isolated S Series Analog Input Block Diagram  
© National Instruments Corporation  
4-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 4  
Analog Input  
On S Series devices, each channel uses its own instrumentation amplifier,  
FIFO, multiplexer (mux), and A/D converter (ADC) to achieve  
simultaneous data acquisition. The main blocks featured in the S Series  
analog input circuitry are as follows:  
Mux—By default, the mux is set to route AI signals to the analog front  
end. When you calibrate your device, the state of the mux switches.  
You can manually switch the state of the mux to measure AI GND.  
Instrumentation Amplifier—The instrumentation amplifier can  
amplify or attenuate an AI signal to ensure that you get the maximum  
resolution of the ADC. Some S Series devices provide programmable  
instrumentation amplifiers that allow you to select the input range.  
Analog Trigger(NI 6124 Only) For information about the trigger  
circuitry of S Series devices, refer to the Analog Input Triggering  
section.  
Filter—The filter on these S Series devices minimizes high frequency  
noise and some attenuating signals by 3 dB at 2 MHz.  
ADC—The analog-to-digital converter (ADC) digitizes the AI signal  
by converting the analog voltage into a digital number.  
AI Timing Signals—For information about the analog input timing  
Isolation Barrier and Digital Isolators(NI 6154 Only) The digital  
isolators across the isolation barrier provide a ground break between  
the isolated analog front end and the chassis ground. For more  
information about isolation and digital isolators, refer to the NI 6154  
Isolation and Digital Isolators section of Appendix A, Device-Specific  
Information.  
AI FIFO—A large first-in-first-out (FIFO) buffer, located inside the  
FPGA, holds data during A/D conversions to ensure that no data is lost.  
S Series devices can handle multiple A/D conversion operations with  
DMA, interrupts, or programmed I/O.  
S Series devices support only differential (DIFF) input mode. The channels  
on S Series devices are true differential inputs, meaning both positive and  
negative inputs can carry signals of interest. For more information about  
DIFF input, refer to the Connecting Analog Input Signals section, which  
contains diagrams showing the signal paths for DIFF input mode.  
NI 6124/6154 User Manual  
4-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Analog Input  
Caution Exceeding the differential and common-mode input ranges distorts the input  
signals. Exceeding the maximum input voltage rating can damage the device and the  
computer. NI is not liable for any damage resulting from such signal connections. The  
maximum input voltage ratings can be found in the specifications document for each  
S Series device.  
Input Polarity and Range  
Input range refers to the set of input voltages that an analog input channel  
can digitize with the specified accuracy. On some S Series devices, you can  
individually program the input range of each AI channel.  
The input range affects the resolution of the S Series device for an  
AI channel. Resolution refers to the magnitude of one ADC code. For  
example, a 16-bit ADC converts analog inputs into one of 65,536 (=216)  
codes, meaning one of 65,536 possible digital values. These values are  
spread fairly evenly across the input range. So, for an input range of –5 V  
to 5 V, the code width of a 16-bit ADC is:  
5 V (5 V)  
------------------------------- = 153 μV  
216  
S Series devices support bipolar input ranges. A bipolar input range means  
that the input voltage range is between –Vref and Vref.  
The instrumentation amplifier applies a different gain setting to the  
AI signal depending on the input range. Gain refers to the factor by which  
the instrumentation amplifier multiplies (amplifies) the input signal before  
sending it to the ADC.  
On S Series devices with programmable input ranges, choose an input  
range that matches the expected input range of your signal. A large input  
range can accommodate a large signal variation, but reduces the voltage  
resolution. Choosing a smaller input range improves the voltage resolution,  
but may result in the input signal going out of range. For more information  
about programming these settings, refer to the NI-DAQmx Help or the  
LabVIEW Help in version 8.0 or later.  
© National Instruments Corporation  
4-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 4  
Analog Input  
Working Voltage Range  
On most S Series devices, the PGIA operates normally by amplifying  
signals of interest while rejecting common-mode signals under the  
following three conditions:  
The common-mode voltage (Vcm), which is equivalent to subtracting  
AI <0..x> GND from AI <0..x> –, must be less than 10 V. This Vcm  
is a constant for all range selections.  
The signal voltage (Vs), which is equivalent to subtracting AI <0..x> +  
from AI <0..x> –, must be less than or equal to the range selection of  
the given channel. If Vs is greater than the range selected, the signal  
clips and information are lost.  
The total working voltage of the positive input, which is equivalent to  
(Vcm + Vs), or subtracting AI <0..x> GND from AI <0..x> +, must be  
less than 11 V.  
If any of these conditions are exceeded, the input voltage is clamped until  
the fault condition is removed.  
Note All inputs are protected at up to 35 V.  
(NI 6154 Only) The isolation features of the NI 6154 improve the working  
voltage range in your applications. Refer to the NI 6154 Specifications for  
more information.  
AI Data Acquisition Methods  
When performing analog input measurements, there are several different  
data acquisition methods available. You can either perform software-timed  
or hardware-timed acquisitions:  
Software-Timed Acquisitions—With a software-timed acquisition,  
software controls the rate of the acquisition. Software sends a separate  
command to the hardware to initiate each ADC conversion. In  
NI-DAQmx, software-timed acquisitions are referred to as having On  
Demand timing. Software-timed acquisitions are also referred to as  
immediate or static acquisitions and are typically used for reading a  
single point of data.  
Hardware-Timed Acquisitions—With hardware-timed acquisitions,  
a digital hardware signal controls the rate of the acquisition. This  
signal can be generated internally on your device or provided  
externally.  
NI 6124/6154 User Manual  
4-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 4  
Analog Input  
Hardware-timed acquisitions have several advantages over  
software-timed acquisitions:  
The time between samples can be much shorter.  
The timing between samples can be deterministic.  
Hardware-timed acquisitions can use hardware triggering. For  
more information, refer to Chapter 11, Triggering.  
Hardware-timed operations can be buffered or non-buffered. A buffer  
is a temporary storage in the computer memory where acquired  
samples are stored.  
Buffered—In a buffered acquisition, data is moved from the DAQ  
device’s onboard FIFO memory to a PC buffer using DMA or  
interrupts before it is transferred to ADE memory. Buffered  
acquisitions typically allow for much faster transfer rates than  
non-buffered acquisitions because data is moved in large blocks,  
rather than one point at a time. For more information about DMA  
and interrupts, refer to the Data Transfer Methods section of  
Chapter 10, Bus Interface.  
One property of buffered I/O operations is the sample mode. The  
sample mode can be either finite or continuous.  
Finite sample mode acquisition refers to the acquisitions of a  
specific, predetermined number of data samples. After the  
specified number of samples has been collected into the buffer, the  
acquisition stops. If you use a reference trigger, you must use  
finite sample mode. Refer to the AI Reference Trigger Signal  
section for more information.  
Continuous acquisition refers to the acquisition of an unspecified  
number of samples. Instead of acquiring a set number of data  
samples and stopping, a continuous acquisition continues until  
you stop the operation. A continuous acquisition is also referred  
to as double-buffered or circular-buffered acquisition.  
If data cannot be transferred across the bus fast enough, the data  
in the FIFO will be overwritten and an error will be generated.  
With continuous operations, if the user program does not read data  
out of the PC buffer fast enough to keep up with the data transfer,  
the buffer could reach an overflow condition, causing an error to  
be generated.  
Non-Buffered—In non-buffered acquisitions, data is read directly  
from the FIFO on the device. Typically, hardware-timed  
non-buffered operations are used to read single samples with  
known time increments between them and small latency.  
© National Instruments Corporation  
4-5  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4  
Analog Input  
Analog Input Triggering  
An analog or digital hardware trigger can initiate these actions. All S Series  
devices support digital triggering, and some also support analog triggering.  
To find your device’s triggering options, refer to the specifications  
document for your device.  
The AI Start Trigger Signal and AI Reference Trigger Signal sections  
contain information about the analog input trigger signals.  
Refer to Chapter 11, Triggering, for more information about triggers.  
Connecting Analog Input Signals  
Table 4-1 summarizes the recommended input configuration for different  
types of signal sources for S Series devices.  
Table 4-1. S Series Analog Input Signal Configuration  
Floating Signal Sources  
(Not Connected to Earth Ground)  
Ground-Referenced Signal Sources  
Examples:  
Example:  
Ungrounded thermocouples  
Plug-in instruments with  
non-isolated outputs  
Signal conditioning with isolated  
outputs  
Input  
NI 6124  
Non-Isolated  
Differential  
(DIFF)  
Battery devices  
AI 0+  
AI 0+  
+
+
+
+
V
V
1
1
AI 0–  
AI 0–  
R
AI GND  
AI GND  
NI 6154  
Isolated  
Differential  
(DIFF)  
AI+  
AI+  
AI–  
Isolation  
Barrier  
Isolation  
Barrier  
+
+
+
+
A I–  
NI 6124/6154 User Manual  
4-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 4  
Analog Input  
Refer to the Analog Input Terminal Configuration section for descriptions  
of the input modes.  
Types of Signal Sources  
When configuring the input channels and making signal connections,  
first determine whether the signal sources are floating or ground-referenced:  
Floating Signal Sources—A floating signal source is not connected in  
any way to the building ground system, and instead has an isolated  
ground-reference point. Some examples of floating signal sources are  
outputs of transformers, thermocouples, battery-powered devices,  
optical isolators, and isolation amplifiers. An instrument or device that  
has an isolated output is a floating signal source. You must connect the  
ground reference of a floating signal to the AI ground of the device to  
establish a local or onboard reference for the signal. Otherwise, the  
measured input signal varies as the source floats outside the  
common-mode input range.  
Ground-Referenced Signal Sources—A ground-referenced signal  
source is connected in some way to the building system ground and is,  
therefore, already connected to a common ground point with respect to  
the device, assuming that the computer is plugged into the same power  
system as the source. Non-isolated outputs of instruments and devices  
that plug into the building power system fall into this category.  
The difference in ground potential between two instruments connected  
to the same building power system is typically between 1 mV and  
100 mV, but the difference can be much higher if power distribution  
circuits are improperly connected. If a grounded signal source is  
incorrectly measured, this difference can appear as measurement error.  
Follow the connection instructions for grounded signal sources to  
eliminate this ground potential difference from the measured signal.  
Isolated devices have isolated front ends that are isolated from  
ground-reference signal sources and are not connected to building  
system grounds. Isolated devices require the user to provide a  
ground-reference terminal to which its input signals are referenced.  
Differential Connections for Ground-Referenced Signal Sources  
Figure 4-6 shows how to connect a ground-referenced signal source to a  
channel on a non-isolated S Series device.  
© National Instruments Corporation  
4-7  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Analog Input  
Non-Isolated S Series Device  
Instrumentation  
AI 0+  
Amplifier  
+
Ground-  
Referenced  
+
V
s
Signal  
+
Source  
+
AI 0–  
Measured  
Voltage  
V
m
Common-  
Mode Noise  
and Ground  
Potential  
V
cm  
AI 0 GND  
I/O Connector  
AI 0 Connections Shown  
Figure 4-3. Differential Connection for Ground-Referenced Signals  
on Non-Isolated Devices  
Figure 4-4 shows how to connect a ground-referenced signal source to a  
channel on an isolated S Series device.  
Isolated S Series Device  
Isolation  
Barrier  
Instrumentation  
AI 0+  
Amplifier  
+
Ground-  
Referenced  
Signal  
+
V
s
+
Digital  
Isolators  
Source  
+
AI 0–  
Measured  
V
m
Voltage  
Common-  
Mode Noise  
and Ground  
Potential  
V
cm  
I/O Connector  
AI 0 Connections Shown  
Figure 4-4. Differential Connection for Ground-Referenced Signals  
on Isolated Devices  
NI 6124/6154 User Manual  
4-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 4  
Analog Input  
With these types of connections, the instrumentation amplifier rejects both  
the common-mode noise in the signal and the ground potential difference  
between the signal source and the device ground, shown as Vcm in these  
figures.  
Common-Mode Signal Rejection Considerations  
The instrumentation amplifier can reject any voltage caused by ground  
potential differences between the signal source and the device. In addition,  
the instrumentation amplifier can reject common-mode noise pickup in the  
leads connecting the signal sources to the device. The instrumentation  
amplifier can reject common-mode signals as long as V+in and V–in (input  
signals) are both within the working voltage range of the device.  
Differential Connections for Non-Referenced or Floating Signal Sources  
Figure 4-5 shows how to connect a floating signal source to a channel on a  
non-isolated S Series device.  
Non-Isolated S Series Device  
AI 0+  
Instrumentation  
+
Floating  
Signal  
Source  
Amplifier  
+
V
s
+
AI 0–  
Measured  
Voltage  
V
m
Bias  
Current  
Return  
Paths  
Bias  
Resistor  
AI 0 GND  
I/O Connector  
AI 0 Connections Shown  
Figure 4-5. Differential Connection for Non-Referenced Signals on  
Non-Isolated Devices  
© National Instruments Corporation  
4-9  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 4  
Analog Input  
Figure 4-5 shows a bias resistor connected between AI 0 – and the floating  
signal source ground. This resistor provides a return path for the bias  
current. A value of 10 kΩ to 100 kΩ is usually sufficient. If you do not use  
the resistor and the source is truly floating, the source is not likely to remain  
within the common-mode signal range of the instrumentation amplifier, so  
the instrumentation amplifier saturates, causing erroneous readings. You  
must reference the source to the respective channel ground.  
Figure 4-6 shows how to connect a floating signal source to a channel on  
an isolated S Series device.  
Isolated S Series Device  
Isolation  
Barrier  
AI 0+  
Instrumentation  
+
Amplifier  
Floating  
Signal  
Source  
+
Digital  
Isolators  
V
s
+
AI 0–  
Measured  
V
m
Voltage  
AI 0 Connections Shown  
Figure 4-6. Differential Connection for Non-Referenced Signals on Isolated Devices  
DC-Coupled  
You can connect low source impedance and high source impedance  
DC-coupled sources:  
Low Source ImpedanceYou must reference the source to AI GND.  
The easiest way to make this reference is to connect the positive side  
of the signal to the positive input of the instrumentation amplifier and  
connect the negative side of the signal to AI GND as well as to the  
negative input of the instrumentation amplifier, without using resistors.  
This connection works well for DC-coupled sources with low source  
impedance (less than 100 Ω).  
NI 6124/6154 User Manual  
4-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 4  
Analog Input  
High Source Impedance—For larger source impedances, this  
connection leaves the DIFF signal path significantly off balance. Noise  
that couples electrostatically onto the positive line does not couple  
onto the negative line because it is connected to ground. Hence, this  
noise appears as a DIFF-mode signal instead of a common-mode  
signal, and the instrumentation amplifier does not reject it. In this case,  
instead of directly connecting the negative line to AI GND, connect the  
negative line to AI GND through a resistor that is about 100 times the  
equivalent source impedance. The resistor puts the signal path nearly  
in balance, so that about the same amount of noise couples onto both  
connections, yielding better rejection of electrostatically coupled  
noise. This configuration does not load down the source (other than the  
very high input impedance of the instrumentation amplifier).  
You can fully balance the signal path by connecting another resistor of  
the same value between the positive input and AI GND. This fully  
balanced configuration offers slightly better noise rejection but has the  
disadvantage of loading the source down with the series combination  
(sum) of the two resistors. If, for example, the source impedance is  
2 kΩ and each of the two resistors is 100 kΩ, the resistors load down  
the source with 200 kΩ and produce a –1% gain error.  
AC-Coupled  
Both inputs of the instrumentation amplifier require a DC path to ground in  
order for the instrumentation amplifier to work. If the source is AC-coupled  
(capacitively coupled), the instrumentation amplifier needs a resistor  
between the positive input and AI GND. If the source has low-impedance,  
choose a resistor that is large enough not to significantly load the source but  
small enough not to produce significant input offset voltage as a result of  
input bias current (typically 100 kΩ to 1 MΩ). In this case, connect the  
negative input directly to AI GND. If the source has high output  
impedance, balance the signal path as previously described using the same  
value resistor on both the positive and negative inputs; be aware that there  
is some gain error from loading down the source.  
Field Wiring Considerations  
Environmental noise can seriously affect the measurement accuracy of the  
S Series device if you do not take proper care when running signal wires  
between signal sources and the device. The following recommendations  
apply mainly to AI signal routing, although they also apply to signal  
routing in general.  
© National Instruments Corporation  
4-11  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 4  
Analog Input  
Minimize noise pickup and maximize measurement accuracy by taking the  
following precautions.  
Use differential AI connections to reject common-mode noise.  
Use individually shielded, twisted-pair wires to connect AI signals to  
the device. With this type of wire, the signals attached to the AI + and  
AI – inputs are twisted together and then covered with a shield. You  
then connect this shield only at one point to the signal source ground.  
This kind of connection is required for signals traveling through areas  
with large magnetic fields or high electromagnetic interference.  
Route signals to the device carefully. Keep cabling away from noise  
sources. The most common noise source in a PCI DAQ system is the  
video monitor. Separate the monitor from the analog signals as far as  
possible.  
Separate the signal lines of the S Series device from high-current or  
high-voltage lines. These lines can induce currents in or voltages on  
the signal lines of the S Series device if they run in close parallel paths.  
To reduce the magnetic coupling between lines, separate them by a  
reasonable distance if they run in parallel, or run the lines at right  
angles to each other.  
Do not run signal lines through conduits that also contain power lines.  
Protect signal lines from magnetic fields caused by electric motors,  
welding equipment, breakers, or transformers by running them through  
special metal conduits.  
Refer to the NI Developer Zone document, Field Wiring and Noise  
Considerations for Analog Signals, for more information.  
Minimizing Drift in Differential Mode  
If the readings from the DAQ device are random and drift rapidly, you  
referenced to a level that is considered floating with reference to the device  
ground reference. Even though you are in DIFF mode, you must still  
reference the signal to the same ground level as the device reference. There  
are various methods of achieving this reference while maintaining a high  
common-mode rejection ratio (CMRR). These methods are outlined in the  
Connecting Analog Input Signals section.  
AI GND is an AI common signal that routes directly to the ground  
connection point on the devices. You can use this signal if you need a  
general analog ground connection point to the device.  
NI 6124/6154 User Manual  
4-12  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 4  
Analog Input  
Analog Input Timing Signals  
An acquisition with posttrigger data allows you to view data that is  
acquired after a trigger event is received. A typical posttrigger DAQ  
sequence is shown in Figure 4-7. The sample counter is loaded with the  
specified number of posttrigger samples, in this example, five. The value  
decrements with each pulse on AI Sample Clock (ai/SampleClock), until  
the value reaches zero and all desired samples have been acquired.  
AI Start Trigger  
AI Sample Clock  
Sample Counter  
4
3
2
1
0
Figure 4-7. Typical Posttriggered DAQ Sequence  
An acquisition with pretrigger data allows you to view data that is acquired  
before the trigger of interest, in addition to data acquired after the trigger.  
Figure 4-8 shows a typical pretrigger DAQ sequence. The AI Start Trigger  
signal (ai/StartTrigger) can be either a hardware or software signal. If  
AI Start Trigger is set up to be a software start trigger, an output pulse  
appears on the AI START TRIG line when the acquisition begins. When the  
AI Start Trigger pulse occurs, the sample counter is loaded with the number  
of pretrigger samples, in this example, four. The value decrements with  
each pulse on AI Sample Clock, until the value reaches zero. The sample  
counter is then loaded with the number of posttrigger samples, in this  
example, three.  
AI Start Trigger  
Don't Care  
AI Reference Trigger  
AI Sample Clock  
Sample Counter  
3
2
1
0
2
2
2
1
0
Figure 4-8. Typical Pretriggered DAQ Sequence  
© National Instruments Corporation  
4-13  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Analog Input  
If an AI Reference Trigger (ai/ReferenceTrigger) pulse occurs before the  
specified number of pretrigger samples are acquired, the trigger pulse is  
ignored. Otherwise, when the AI Reference Trigger pulse occurs, the  
sample counter value decrements until the specified number of posttrigger  
samples have been acquired. For more information about start and  
reference triggers, refer to the Analog Input Triggering section.  
In order to provide all of the timing functionality described throughout this  
section, the DAQ-STC2 provides an extremely powerful and flexible  
timing engine. For more information about all of the clock routing and  
timing options that the analog input timing engine provides, refer to the  
NI-DAQmx Help or the LabVIEW Help in version 8.0 or later.  
S Series devices feature the following analog input timing signals:  
AI Sample Clock Signal  
AI Sample Clock Timebase Signal  
AI Convert Clock Signal  
AI Convert Clock Timebase Signal  
AI Hold Complete Event Signal  
AI Start Trigger Signal  
AI Reference Trigger Signal  
AI Sample Clock Signal  
Use the AI Sample Clock (ai/SampleClock) signal to initiate a set of  
measurements. Your S Series device samples the AI signals of every  
channel in the task once for every AI Sample Clock.  
You can specify an internal or external source for AI Sample Clock. You  
also can specify whether the measurement sample begins on the rising edge  
or falling edge of AI Sample Clock.  
Using an Internal Source  
One of the following internal signals can drive AI Sample Clock:  
Counter n Internal Output  
AI Sample Clock Timebase (divided down)  
A pulse initiated by host software  
A programmable internal counter divides down the sample clock timebase.  
NI 6124/6154 User Manual  
4-14  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 4  
Analog Input  
Several other internal signals can be routed to AI Sample Clock through  
RTSI. Refer to Device Routing in MAX in the NI-DAQmx Help or the  
LabVIEW Help in version 8.0 or later for more information.  
Using an External Source  
Use one of the following external signals as the source of AI Sample Clock:  
PFI <0..15>  
RTSI <0..7>  
PXI_STAR  
Analog Comparison Event (an analog trigger)  
Routing AI Sample Clock Signal to an Output  
Terminal  
You can route AI Sample Clock out to any PFI <0..15> or RTSI <0..7>  
terminal. This pulse is always active high.  
You can specify the output to have one of two behaviors. With the pulse  
behavior, your DAQ device briefly pulses the PFI terminal once for every  
occurrence of AI Sample Clock.  
With level behavior, your DAQ device drives the PFI terminal high during  
the entire sample.  
All PFI terminals are configured as inputs by default.  
Other Timing Requirements  
A counter on your device internally generates AI Sample Clock unless you  
select some external source. AI Start Trigger starts this counter and either  
software or hardware can stop it once a finite acquisition completes. When  
using an internally generated AI Sample Clock, you also can specify a  
configurable delay from AI Start Trigger to the first AI Sample Clock  
pulse. By default, this delay is set to two ticks of the AI Sample Clock  
Timebase signal.  
© National Instruments Corporation  
4-15  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 4  
Analog Input  
Figure 4-9 shows the relationship of AI Sample Clock to AI Start Trigger.  
AI Sample Clock Timebase  
AI Start Trigger  
AI Sample Clock  
Delay  
From  
Start  
Trigger  
Figure 4-9. AI Sample Clock and AI Start Trigger  
AI Sample Clock Timebase Signal  
You can route any of the following signals to be the AI Sample Clock  
Timebase (ai/SampleClockTimebase) signal:  
20 MHz Timebase  
100 kHz Timebase  
PXI_CLK10  
RTSI <0..7>  
PFI <0..15>  
PXI_STAR  
Analog Comparison Event (an analog trigger)  
AI Sample Clock Timebase is not available as an output on the  
I/O connector. AI Sample Clock Timebase is divided down to provide one  
of the possible sources for AI Sample Clock. You can configure the polarity  
selection for AI Sample Clock Timebase as either rising or falling edge.  
AI Convert Clock Signal  
Use the AI Convert Clock (ai/ConvertClock) signal to initiate a single A/D  
conversion on every channel.  
You can specify either an internal or external signal as the source of  
AI Convert Clock. You also can specify whether the measurement sample  
begins on the rising edge or falling edge of AI Convert Clock.  
NI 6124/6154 User Manual  
4-16  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 4  
Analog Input  
Using an Internal Source  
One of the following internal signals can drive AI Convert Clock:  
AI Convert Clock Timebase (divided down)  
Counter n Internal Output  
A programmable internal counter divides down the AI Convert Clock  
Timebase to generate AI Convert Clock. The counter is started by  
AI Sample Clock and continues to count down to zero, produces an  
AI Convert Clock, reloads itself, and repeats the process until the sample is  
finished. It then reloads itself in preparation for the next AI Sample Clock  
pulse.  
Using an External Source  
Use one of the following external signals as the source of AI Convert  
Clock:  
PFI <0..15>  
RTSI <0..7>  
PXI_STAR  
Analog Comparison Event (an analog trigger)  
Routing AI Convert Clock Signal to an Output  
Terminal  
You can route AI Convert Clock (as an active low signal) out to any  
PFI <0..15> or RTSI <0..7> terminal.  
All PFI terminals are configured as inputs by default.  
AI Convert Clock Timebase Signal  
The AI Convert Clock Timebase (ai/ConvertClockTimebase) signal is  
divided down to provide on of the possible sources for AI Convert Clock.  
Use one of the following signals as the source of AI Convert Clock  
Timebase:  
AI Sample Clock Timebase  
20 MHz Timebase  
AI Convert Clock Timebase is not available as an output on the I/O  
connector.  
© National Instruments Corporation  
4-17  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 4  
Analog Input  
AI Hold Complete Event Signal  
The AI Hold Complete Event (ai/HoldCompleteEvent) signal  
generates a pulse after each A/D conversion begins. You can route  
ai/HoldCompleteEvent out to any PFI <0..15> or RTSI <0..7> terminal.  
The polarity of ai/HoldCompleteEvent is software-selectable, but is  
typically configured so that a low-to-high leading edge can clock external  
AI multiplexers indicating when the input signal has been sampled and can  
be removed.  
AI Start Trigger Signal  
Use the AI Start Trigger (ai/StartTrigger) signal to begin a measurement  
acquisition. A measurement acquisition consists of one or more samples.  
If you do not use triggers, begin a measurement with a software command.  
Once the acquisition begins, configure the acquisition to stop:  
When a certain number of points are sampled (in finite mode)  
After a hardware reference trigger (in finite mode)  
With a software command (in continuous mode)  
An acquisition that uses a start trigger (but not a reference trigger) is  
sometimes referred to as a posttriggered acquisition.  
Using a Digital Source  
To use AI Start Trigger with a digital source, specify a source and an edge.  
The source can be any of the following signals:  
PFI <0..15>  
RTSI <0..7>  
Counter n Internal Output  
PXI_STAR  
The source also can be one of several other internal signals on your DAQ  
device. Refer to Device Routing in MAX in the NI-DAQmx Help or the  
LabVIEW Help in version 8.0 or later for more information.  
You also can specify whether the measurement acquisition begins on the  
rising edge or falling edge of AI Start Trigger.  
NI 6124/6154 User Manual  
4-18  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Analog Input  
Using an Analog Source  
When you use an analog trigger source, the acquisition begins on the  
first rising edge of the Analog Comparison Event signal.  
Routing AI Start Trigger to an Output Terminal  
You can route AI Start Trigger out to any PFI <0..15> or RTSI <0..7>  
terminal. The output is an active high pulse. All PFI terminals are  
configured as inputs by default.  
The device also uses AI Start Trigger to initiate pretriggered DAQ  
operations. In most pretriggered applications, a software trigger generates  
AI Start Trigger. Refer to the AI Reference Trigger Signal section for a  
complete description of the use of AI Start Trigger and AI Reference  
Trigger in a pretriggered DAQ operation.  
AI Reference Trigger Signal  
Use AI Reference Trigger (ai/ReferenceTrigger) signal to stop a  
measurement acquisition. To use a reference trigger, specify a buffer of  
finite size and a number of pretrigger samples (samples that occur before  
the reference trigger). The number of posttrigger samples (samples that  
occur after the reference trigger) desired is the buffer size minus the  
number of pretrigger samples.  
Once the acquisition begins, the DAQ device writes samples to the buffer.  
After the DAQ device captures the specified number of pretrigger samples,  
the DAQ device begins to look for the reference trigger condition. If the  
reference trigger condition occurs before the DAQ device captures the  
specified number of pretrigger samples, the DAQ device ignores the  
condition.  
If the buffer becomes full, the DAQ device continuously discards the oldest  
samples in the buffer to make space for the next sample. This data can be  
accessed (with some limitations) before the DAQ device discards it. Refer  
to the KnowledgeBase document, Can a Pretriggered Acquisition be  
Continuous?, for more information. To access this KnowledgeBase, go to  
ni.com/infoand enter the info code rdcanq.  
© National Instruments Corporation  
4-19  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Analog Input  
When the reference trigger occurs, the DAQ device continues to write  
samples to the buffer until the buffer contains the number of posttrigger  
samples desired. Figure 4-10 shows the final buffer.  
Reference Trigger  
Pretrigger Samples  
Complete Buffer  
Figure 4-10. Reference Trigger Final Buffer  
Using a Digital Source  
To use AI Reference Trigger with a digital source, specify a source and an  
edge. The source can be any of the following signals:  
PFI <0..15>  
RTSI <0..7>  
PXI_STAR  
The source also can be one of several internal signals on your DAQ device.  
Refer to Device Routing in MAX in the NI-DAQmx Help or the LabVIEW  
Help in version 8.0 or later for more information.  
You also can specify whether the measurement acquisition stops on the  
rising edge or falling edge of AI Reference Trigger.  
Using an Analog Source  
When you use an analog trigger source, the acquisition stops on the  
first rising edge of the Analog Comparison Event signal.  
Routing AI Reference Trigger Signal to an Output  
Terminal  
You can route AI Reference Trigger out to any PFI <0..15> or RTSI <0..7>  
terminal.  
All PFI terminals are configured as inputs by default.  
NI 6124/6154 User Manual  
4-20  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 4  
Analog Input  
Getting Started with AI Applications in Software  
You can use the S Series device in the following analog input applications:  
Simultaneous sampling  
Single-point analog input  
Finite analog input  
Continuous analog input  
You can perform these applications through DMA, interrupt, or  
programmed I/O data transfer mechanisms. Some of the applications also  
use start and reference pause triggers.  
Note For more information about programming analog input applications and triggers in  
software, refer to the NI-DAQmx Help or the LabVIEW Help in version 8.0 or later.  
© National Instruments Corporation  
4-21  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
5
Analog Output  
Figure 5-1 shows the analog output circuitry of a non-isolated S Series  
(NI 6124) device.  
DAC0  
DAC1  
AO 0  
AO 1  
AO FIFO  
AO Data  
AO Sample Clock  
Figure 5-1. Non-Isolated S Series Device Analog Output Block Diagram  
Figure 5-2 shows the analog output circuitry of an isolated S Series  
(NI 6154) device.  
Isolation  
Barrier  
AO+  
DAC0  
AO Data  
AO FIFO  
Digital  
Isolators  
AO Sample Clock  
Figure 5-2. Isolated S Series Device Analog Output Block Diagram  
© National Instruments Corporation  
5-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 5  
Analog Output  
The main blocks featured in the S Series analog output circuitry are as  
follows:  
AO FIFO—The AO FIFO enables analog output waveform  
generation. It is a first-in-first-out (FIFO) memory buffer between the  
computer and the DACs that allows you to download all the points of  
a waveform to your board without host computer interaction.  
AO Sample Clock—The DAC reads a sample from the FIFO with  
every cycle of the AO Sample Clock signal and generates the AO  
Isolation Barrier and Digital Isolators(NI 6154 Only) The digital  
isolators across the isolation barrier provide a ground break between  
the isolated analog front end and the chassis ground. For more  
information about isolation and digital isolators, refer to the NI 6154  
Isolation and Digital Isolators section of Appendix A, Device-Specific  
Information.  
DAC—Digital-to-analog converters (DACs) convert digital codes to  
analog voltages.  
Minimizing Glitches on the Output Signal  
When you use a DAC to generate a waveform, you may observe glitches  
on the output signal. These glitches are normal; when a DAC switches from  
one voltage to another, it produces glitches due to released charges. The  
largest glitches occur when the most significant bit (MSB) of the DAC code  
switches. You can build a lowpass deglitching filter to remove some of  
these glitches, depending on the frequency and nature of the output signal.  
Visit ni.com/supportfor more information about minimizing glitches.  
AO Data Generation Methods  
When performing an analog output operation, there are several different  
data generation methods available. You can either perform software-timed  
or hardware-timed generations:  
Software-Timed Generations—With a software-timed generation,  
software controls the rate at which data is generated. Software sends a  
separate command to the hardware to initiate each DAC conversion. In  
NI-DAQmx, software-timed generations are referred to as On Demand  
timing. Software-timed generations are also referred to as immediate  
or static operations. They are typically used for writing a single value  
out, such as a constant DC voltage.  
NI 6124/6154 User Manual  
5-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 5  
Analog Output  
Hardware-Timed Generations—With a hardware-timed generation,  
a digital hardware signal controls the rate of the generation. This signal  
can be generated internally on your device or provided externally.  
Hardware-timed generations have several advantages over  
software-timed generations:  
The time between samples can be much shorter.  
The timing between samples can be deterministic.  
Hardware-timed generations can use hardware triggering. For  
more information, refer to Chapter 11, Triggering.  
Hardware-timed operations can be buffered or non-buffered. A buffer  
is a temporary storage in computer memory for acquired or  
to-be-generated samples.  
Buffered—In a buffered generation, data is moved from a PC  
buffer to the DAQ device’s onboard FIFO using DMA or  
interrupts before it is written to the DACs one sample at a time.  
Buffered generations typically allow for much faster transfer rates  
than non-buffered generations because data is moved in large  
blocks, rather than one point at a time. For more information about  
DMA and interrupts, refer to the Data Transfer Methods section  
of Chapter 10, Bus Interface.  
One property of buffered I/O operations is the sample mode. The  
sample mode can be either finite or continuous.  
Finite sample mode generation refers to the generation of a  
specific, predetermined number of data samples. When the  
specified number of samples has been written out, the generation  
stops.  
Continuous generation refers to the generation of an unspecified  
number of samples. Instead of generating a set number of data  
samples and stopping, a continuous generation continues until  
you stop the operation. There are several different methods of  
continuous generation that control what data is written. These  
methods are regeneration, FIFO regeneration and  
non-regeneration modes.  
Regeneration is the repetition of the data that is already in the  
buffer. Standard regeneration is when data from the PC buffer is  
continually downloaded to the FIFO to be written out. New data  
can be written to the PC buffer at any time without disrupting the  
output.  
© National Instruments Corporation  
5-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5  
Analog Output  
With FIFO regeneration, the entire buffer is downloaded to the  
FIFO and regenerated from there. After the data is downloaded,  
new data cannot be written to the FIFO. To use FIFO regeneration,  
the entire buffer must fit within the FIFO size. The advantage of  
using FIFO regeneration is that it does not require communication  
with the main host memory after the operation is started, thereby  
preventing any problems that may occur due to excessive bus  
traffic.  
With non-regeneration, old data will not be repeated. New data  
must be continually written to the buffer. If the program does not  
write new data to the buffer at a fast enough rate to keep up with  
the generation, the buffer will underflow and cause an error.  
Non-Buffered—In hardware-timed non-buffered generations,  
data is written directly to the FIFO on the device. Typically,  
hardware-timed non-buffered operations are used to write single  
samples with known time increments between them and good  
latency.  
Analog Output Triggering  
analog or digital hardware trigger can initiate these actions. All S Series  
devices support digital triggering, and some also support analog triggering.  
To find your device’s triggering options, refer to the specifications  
document for your device.  
The AO Start Trigger Signal and AO Pause Trigger Signal sections contain  
information about the analog output trigger signals.  
Refer to Chapter 11, Triggering, for more information about triggers.  
Connecting Analog Output Signals  
The AO signals are AO 0, AO 1, and AO GND. AO 0 is the voltage output  
signal for AO channel 0. AO 1 is the voltage output signal for  
AO channel 1. AO GND is the ground reference for the AO channels.  
NI 6124/6154 User Manual  
5-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 5  
Analog Output  
Figure 5-3 shows how AO 0 and AO 1 are wired on a non-isolated S Series  
device.  
AO 0  
Channel 0  
+
Load  
VOUT 0  
AO GND  
AO 1  
Load  
VOUT 1  
+
Channel 1  
Analog Output Channels  
Non-Isolated S Series Device  
Figure 5-3. Analog Output Connections for Non-Isolated S Series Devices  
Figure 5-4 shows how AO 0 is wired on an isolated S Series device.  
Isolation  
Barrier  
AO+  
AO–  
DAC  
+
Digital  
Isolators  
Analog Output Channel  
Load  
VOUT  
Isolated S Series Device  
Figure 5-4. Analog Output Connections for Isolated S Series Devices  
© National Instruments Corporation  
5-5  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 5  
Analog Output  
Waveform Generation Timing Signals  
There is one AO Sample Clock that causes all AO channels to update  
simultaneously. Figure 5-5 summarizes the timing and routing options  
provided by the analog output timing engine.  
Ctr1InternalOutput  
PFI 0–9,  
RTSI 0–6  
RTSI 7  
Master  
Timebase  
PFI 0–9,  
ao/SampleClock  
Onboard  
Clock  
ao/SampleClock  
RTSI 0–6  
Onboard  
Timebase  
Clock  
÷
÷200  
20 MHz  
Divisor  
Timebase  
S Series devices feature the following waveform generation timing signals:  
AO Sample Clock Signal  
AO Sample Clock Timebase Signal  
AO Start Trigger Signal  
AO Pause Trigger Signal  
AO Sample Clock Signal  
You can use the AO Sample Clock (ao/SampleClock) signal to initiate  
AO samples. Each sample updates the outputs of all of the DACs.  
The source of the AO Sample Clock signal can be internal or external. You  
can specify whether the DAC update begins on the rising edge or falling  
edge of the AO Sample Clock signal.  
Using an Internal Source  
By default, AO Sample Clock is created internally by dividing down the  
AO Sample Clock Timebase signal.  
Several other internal signals can be routed to the sample clock. Refer to  
Device Routing in MAX in the NI-DAQmx Help or the LabVIEW Help in  
version 8.0 or later for more information.  
NI 6124/6154 User Manual  
5-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 5  
Analog Output  
Using an External Source  
You can use a signal connected to any PFI or RTSI <0..6> pin as the source  
of AO Sample Clock. Figure 5-6 shows the timing requirements of the  
AO Sample Clock source.  
t
w
Rising-Edge  
Polarity  
Falling-Edge  
Polarity  
t
= 10 ns minimum  
w
Figure 5-6. AO Sample Clock Timing Requirements  
Routing AO Sample Clock Signal to an Output  
Terminal  
You can route ao/SampleClock (as an active low signal) out to any  
PFI <0..15> or RTSI <0..7> terminal.  
Other Timing Requirements  
A counter on your device internally generates AO Sample Clock unless you  
select some external source. The AO Start Trigger signal starts this counter.  
It is stopped automatically by hardware after a finite acquisition completes  
or manually through software. When using an internally generated  
AO Sample Clock in NI-DAQmx, you can also specify a configurable  
delay from the AO Start Trigger to the first AO Sample Clock pulse. By  
default, this delay is two ticks of the AO Sample Clock Timebase signal.  
© National Instruments Corporation  
5-7  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 5  
Analog Output  
Figure 5-7 shows the relationship of the AO Sample Clock signal to the  
AO Start Trigger signal.  
AO Sample Clock Timebase  
AO Start Trigger  
AO Sample Clock  
Delay  
From  
Start  
Trigger  
Figure 5-7. AO Sample Clock and AO Start Trigger  
AO Sample Clock Timebase Signal  
You can select any PFI or RTSI pin as well as many other internal signals  
as the AO Sample Clock Timebase (ao/SampleClockTimebase) signal.  
This signal is not available as an output on the I/O connector. AO Sample  
Clock Timebase is divided down to provide the Onboard Clock source for  
the AO Sample Clock. You specify whether the samples begin on the rising  
or falling edge of AO Sample Clock Timebase.  
You might use the AO Sample Clock Timebase signal if you want to use an  
external sample clock signal, but need to divide the signal down. If you  
want to use an external sample clock signal, but do not need to divide the  
signal, then you should use the AO Sample Clock signal rather than the  
AO Sample Clock Timebase. If you do not specify an external sample clock  
timebase, NI-DAQmx uses the Onboard Clock.  
NI 6124/6154 User Manual  
5-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 5  
Analog Output  
Figure 5-8 shows the timing requirements for the AO Sample Clock  
Timebase signal.  
tp  
tw  
tw  
tp = 50 ns minimum  
t
w = 23 ns minimum  
Figure 5-8. AO Sample Clock Timebase Timing Requirements  
The maximum allowed frequency is 20 MHz, with a minimum pulse width  
of 10 ns high or low. There is no minimum frequency.  
Unless you select an external source, either the 20MHzTimebase or  
100kHzTimebase generates the AO Sample Clock Timebase signal.  
AO Start Trigger Signal  
You can use the AO Start Trigger signal (ao/StartTrigger) to initiate a  
waveform generation. If you do not use triggers, you begin a generation  
with a software command.  
Using a Digital Source  
To use AO Start Trigger, specify a source and an edge. The source can be  
an external signal connected to any PFI or RTSI <0..6> pin. The source can  
also be one of several internal signals on your DAQ device. Refer to Device  
Routing in MAX in the NI-DAQmx Help or the LabVIEW Help in  
version 8.0 or later for more information.  
© National Instruments Corporation  
5-9  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 5  
Analog Output  
Figure 5-9 shows the timing requirements of the AO Start Trigger digital  
source.  
tw  
Rising-Edge  
Polarity  
Falling-Edge  
Polarity  
tw = 10 ns minimum  
Figure 5-9. AO Start Trigger Timing Requirements  
Using an Analog Source  
When you use an analog trigger source, the waveform generation begins on  
the first rising edge of the Analog Comparison Event signal. For more  
information, refer to the Triggering with an Analog Source section of  
Chapter 11, Triggering.  
Routing AO Start Trigger Signal to an Output  
Terminal  
You can route ao/StartTrigger out to any PFI <0..15> or RTSI <0..7>  
terminal.  
The output is an active high pulse.  
PFI terminals are configured as inputs by default.  
AO Pause Trigger Signal  
You can use the AO Pause Trigger signal (ao/PauseTrigger) to mask off  
samples in a DAQ sequence. That is, when AO Pause Trigger is active,  
no samples occur.  
The AO Pause Trigger does not stop a sample that is in progress. The pause  
does not take effect until the beginning of the next sample. This signal is  
not available as an output.  
NI 6124/6154 User Manual  
5-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 5  
Analog Output  
Using a Digital Source  
To use ao/Pause Trigger, specify a source and a polarity. The source can be  
an external signal connected to any PFI or RTSI <0..6> pin. The source can  
also be one of several other internal signals on your DAQ device. Refer to  
Device Routing in MAX in the NI-DAQmx Help or the LabVIEW Help in  
version 8.0 or later for more information.  
Also, specify whether the samples are paused when AO Pause Trigger is at  
a logic high or low level.  
Using an Analog Source  
When you use an analog trigger source, the samples are paused when the  
Analog Comparison Event signal is at a high level. For more information,  
refer to the Triggering with an Analog Source section of Chapter 11,  
Triggering.  
Getting Started with AO Applications in Software  
You can use the S Series device in the following analog output  
applications:  
Single-point generation  
Finite generation  
Continuous generation  
Waveform generation  
You can perform these generations through DMA, interrupt, or  
programmed I/O data transfer mechanisms. Some of the applications also  
use start triggers and pause triggers.  
Note For more information about programming analog output applications and triggers in  
software, refer to the NI-DAQmx Help or the LabVIEW Help in version 8.0 or later.  
© National Instruments Corporation  
5-11  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
6
Digital I/O  
Refer to one of the following sections, depending on your device:  
Digital I/O for Non-Isolated Devices—NI 6124 devices have eight  
lines of bidirectional DIO lines on Port 0, and 16 PFI signals that can  
function as static DIO lines.  
Digital I/O for Isolated Devices—NI 6154 devices have six  
bank-isolated digital inputs and four bank-isolated digital outputs.  
Digital I/O for Non-Isolated Devices  
(NI 6124 Only) NI 6124 devices contain eight lines of bidirectional DIO lines  
on Port 0. In addition, The NI 6124 has 16 PFI lines that can function as  
static DIO lines.  
These S Series devices support the following DIO features on Port 0:  
Eight lines of DIO  
Direction and function of each terminal individually controllable  
Static digital input and output  
High-speed digital waveform generation  
High-speed digital waveform acquisition  
DI change detection trigger/interrupt  
© National Instruments Corporation  
6-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 6  
Digital I/O  
Figure 6-1 shows the circuitry of one DIO line. Each DIO line is similar.  
The following sections provide information about the various parts of the  
DIO circuit.  
DO Waveform  
Generation FIFO  
DO Sample Clock  
Static DO  
Buffer  
P0.x  
I/O Protection  
DO.x Direction Control  
Weak Pull-Down  
Static DI  
DI Waveform  
Measurement  
FIFO  
DI Sample Clock  
DI Change  
Detection  
Figure 6-1. Non-Isolated S Series Digital I/O Circuitry  
The DIO terminals are named P0.<0..7> on the device I/O connector.  
The voltage input and output levels and the current drive levels of the DIO  
lines are listed in the specifications of your device.  
Static DIO for Non-Isolated Devices  
(NI 6124 Only) Each DIO line can be used as a static DI or DO line. You can  
use static DIO lines to monitor or control digital signals. Each DIO can be  
individually configured as a digital input (DI) or digital output (DO).  
All samples of static DI lines and updates of DO lines are software-timed.  
P0.6 and P0.7 on these devices also can control the up/down input of  
general-purpose counters 0 and 1, respectively. However, it is  
recommended that you use PFI signals to control the up/down input of the  
counters. The up/down control signals, Counter 0 Up_Down and Counter 1  
Up_Down, are input-only and do not affect the operation of the DIO lines.  
NI 6124/6154 User Manual  
6-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 6  
Digital I/O  
Digital Waveform Triggering for Non-Isolated Devices  
(NI 6124 Only) NI 6124 devices do not have an independent DI or DO Start  
Trigger for digital waveform acquisition or generation. To trigger a DI or  
DO operation, first select a signal to be the source of DI Sample Clock or  
DO Sample Clock. Then, generate a trigger that initiates pulses on the  
source signal. The method for generating this trigger depends on which  
signal is the source of DI Sample Clock or DO Sample Clock.  
For example, consider the case where you are using AI Sample Clock as the  
source of DI Sample Clock. To initiate pulses on AI Sample Clock (and  
therefore on DI Sample Clock), you use AI Start Trigger to trigger the start  
of an AI operation. The AI Start Trigger causes the non-isolated  
DAQ-STC2 S Series device to begin generating AI Sample clock pulses,  
which in turn generates DI Sample clock pulses, as shown in Figure 6-2.  
PFI 1  
AI Start Trigger initiates  
(AI Start Trigger)  
AI Sample Clock and DI Sample Clock.  
AI Sample Clock  
DI Sample Clock  
Figure 6-2. Digital Waveform Triggering  
Similarly, if you are using AO Sample Clock as the source of DI Sample  
Clock, then AO Start Trigger initiates both AO and DI operations.  
If you are using a Counter output as the source of DI Sample Clock, the  
counter’s start trigger, enables the counter which drives DI Sample Clock.  
If you are using an external signal (such as PFI x) as the source for  
DI Sample Clock or DO Sample Clock, you must trigger that external  
signal.  
Digital Waveform Acquisition for Non-Isolated Devices  
(NI 6124 Only) You can acquire digital waveforms on the Port 0 DIO lines.  
The DI waveform acquisition FIFO stores the digital samples. S Series  
devices have a DMA controller dedicated to moving data from the  
DI waveform acquisition FIFO to system memory. The DAQ device  
samples the DIO lines on each rising or falling edge of a clock signal,  
DI Sample Clock.  
© National Instruments Corporation  
6-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 6  
Digital I/O  
You can configure each DIO line to be an output, a static input, or a digital  
waveform acquisition input.  
DI Sample Clock Signal  
(NI 6124 Only) Use the DI Sample Clock (di/SampleClock) signal to sample  
the P0.<0..7> terminals and store the result in the DI waveform acquisition  
FIFO. These S Series devices do not have the ability to divide down a  
timebase to produce an internal DI Sample Clock for digital waveform  
acquisition. Therefore, you must route an external signal or one of many  
internal signals from another subsystem to be the DI Sample Clock. For  
example, you can correlate digital and analog samples in time by sharing  
your AI Sample Clock or AO Sample Clock as the source of your  
DI Sample Clock. To sample a digital signal independent of an AI, AO,  
or DO operation, you can configure a counter to generate the desired  
DI Sample Clock or use an external signal as the source of the clock.  
If the DAQ device receives a DI Sample Clock when the FIFO is full,  
it reports an overflow error to the host software.  
Using an Internal Source  
To use DI Sample Clock with an internal source, specify the signal source  
and the polarity of the signal. The source can be any of the following  
signals:  
AI Sample Clock (ai/SampleClock)  
AI Convert Clock (ai/ConvertClock)  
AO Sample Clock (ao/SampleClock)  
Counter n Internal Output  
Frequency Output  
DI Change Detection Output  
Several other internal signals can be routed to DI Sample Clock through  
RTSI. Refer to Device Routing in MAX in the NI-DAQmx Help or the  
LabVIEW Help in version 8.0 or later for more information.  
Using an External Source  
You can route any of the following signals as DI Sample Clock:  
RTSI <0..7>  
NI 6124/6154 User Manual  
6-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 6  
Digital I/O  
PXI_STAR  
Analog Comparison Event (an analog trigger)  
You can sample data on the rising or falling edge of DI Sample Clock.  
Routing DI Sample Clock to an Output Terminal  
You can route DI Sample Clock out to any PFI terminal. The PFI circuitry  
inverts the polarity of DI Sample Clock before driving the PFI terminal.  
Digital Waveform Generation for Non-Isolated Devices  
(NI 6124 Only) You can generate digital waveforms on the Port 0 DIO lines.  
The DO waveform generation FIFO stores the digital samples. These  
S Series devices have a DMA controller dedicated to moving data from the  
system memory to the DO waveform generation FIFO. The DAQ device  
moves samples from the FIFO to the DIO terminals on each rising or falling  
edge of a clock signal, DO Sample Clock. You can configure each DIO  
signal to be an input, a static output, or a digital waveform generation  
output.  
The FIFO supports a retransmit mode. In the retransmit mode, after all the  
samples in the FIFO have been clocked out, the FIFO begins outputting all  
of the samples again in the same order. For example, if the FIFO contains  
five samples, the pattern generated consists of sample #1, #2, #3, #4, #5,  
#1, #2, #3, #4, #5, #1, and so on.  
DO Sample Clock Signal  
(NI 6124 Only) Use the DO Sample Clock (do/SampleClock) signal to update  
the DO terminals with the next sample from the DO waveform generation  
FIFO. These S Series devices do not have the ability to divide down a  
timebase to produce an internal DO Sample Clock for digital waveform  
generation. Therefore, you must route an external signal or one of many  
internal signals from another subsystem to be the DO Sample Clock. For  
example, you can correlate digital and analog samples in time by sharing  
your AI Sample Clock or AO Sample Clock as the source of your  
DO Sample Clock. To generate digital data independent of an AI, AO,  
or DI operation, you can configure a counter to generate the desired  
DO Sample Clock or use an external signal as the source of the clock.  
If the DAQ device receives a DO Sample Clock when the FIFO is empty,  
the DAQ device reports an underflow error to the host software.  
© National Instruments Corporation  
6-5  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 6  
Digital I/O  
Using an Internal Source  
To use DO Sample Clock with an internal source, specify the signal source  
and the polarity of the signal. The source can be any of the following  
signals:  
AI Sample Clock (ai/SampleClock)  
AI Convert Clock (ai/ConvertClock)  
AO Sample Clock (ao/SampleClock)  
Counter n Internal Output  
Frequency Output  
DI Change Detection Output  
Several other internal signals can be routed to DO Sample Clock through  
RTSI. Refer to Device Routing in MAX in the NI-DAQmx Help or the  
LabVIEW Help in version 8.0 or later for more information.  
Using an External Source  
You can route any of the following signals as DO Sample Clock:  
PFI <0..15>  
RTSI <0..7>  
PXI_STAR  
Analog Comparison Event (an analog trigger)  
You can generate samples on the rising or falling edge of DO Sample  
Clock.  
You must ensure that the time between two active edges of DO Sample  
Clock is not too short. If the time is too short, the DO waveform generation  
FIFO is not able to read the next sample fast enough. The DAQ device  
reports an overrun error to the host software.  
Routing DO Sample Clock to an Output Terminal  
You can route DO Sample Clock out to any PFI terminal. The PFI circuitry  
inverts the polarity of DO Sample Clock before driving the PFI terminal.  
NI 6124/6154 User Manual  
6-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6  
Digital I/O  
I/O Protection for Non-Isolated Devices  
(NI 6124 Only) Each DIO and PFI signal is protected against overvoltage,  
undervoltage, and overcurrent conditions as well as ESD events. However,  
you should avoid these fault conditions by following these guidelines:  
If you configure a PFI or DIO line as an output, do not connect it to any  
external signal source, ground, or power supply.  
If you configure a PFI or DIO line as an output, understand the current  
requirements of the load connected to these signals. Do not exceed the  
specified current output limits of the DAQ device. NI has several signal  
conditioning solutions for digital applications requiring high current  
drive.  
If you configure a PFI or DIO line as an input, do not drive the line with  
voltages outside of its normal operating range. The PFI or DIO lines  
have a smaller operating range than the AI signals.  
Treat the DAQ device as you would treat any static sensitive device.  
Always properly ground yourself and the equipment when handling  
the DAQ device or connecting to it.  
Programmable Power-Up States for Non-Isolated Devices  
(NI 6124 Only) At system startup and reset, the hardware sets all PFI and DIO  
lines to high-impedance inputs by default. The DAQ device does not drive  
the signal high or low. Each line has a weak pull-down resistor connected  
to it, as described in the specifications document for your device.  
NI-DAQmx supports programmable power-up states for PFI and DIO lines.  
Software can program any value at power up to the P0, P1, or P2 lines. The  
PFI and DIO lines can be set as:  
A high-impedance input with a weak pull-down resistor (default)  
An output driving a 0  
An output driving a 1  
Refer to the NI-DAQmx Help or the LabVIEW Help in version 8.0 or later  
for more information about setting power-up states in NI-DAQmx or MAX.  
Note When using your S Series device to control an SCXI chassis, DIO lines 0, 1, 2, and  
4 are used as communication lines and must be left to power-up in the default  
high-impedance state to avoid potential damage to these signals.  
© National Instruments Corporation  
6-7  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 6  
Digital I/O  
DI Change Detection for Non-Isolated Devices  
(NI 6124 Only) You can configure the DAQ device to detect changes in the  
DIO signals. Figure 6-3 shows a block diagram of the DIO change  
detection circuitry.  
Enable  
P0.0  
Synch  
Enable  
Change Detection Event  
Enable  
P0.7  
Synch  
Enable  
Figure 6-3. DI Change Detection  
You can enable the DIO change detection circuitry to detect rising edges,  
falling edges, or either edge individually on each DIO line. The DAQ  
devices synchronize each DI signal to 80MHzTimebase, and then sends the  
signal to the change detectors. The circuitry ORs the output of all enabled  
change detectors from every DI signal. The result of this OR is the Change  
Detection Event signal.  
The Change Detection Event signal can do the following:  
Drive any RTSI <0..7>, PFI <0..15>, or PXI_STAR signal  
Drive the DO Sample Clock or DI Sample Clock  
Generate an interrupt  
NI 6124/6154 User Manual  
6-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 6  
Digital I/O  
The Change Detection Event signal also can be used to detect changes on  
digital output events.  
DI Change Detection Applications for Non-Isolated  
Devices  
program when one of several DIO signals changes state.  
You also can use the output of the DIO change detection circuitry to trigger  
a DI or counter acquisition on the logical OR of several digital signals. To  
trigger on a single digital signal, refer to the Triggering with a Digital  
Source section of Chapter 11, Triggering. By routing the Change Detection  
Event signal to a counter, you also can capture the relative time between  
samples.  
You also can use the Change Detection Event signal to trigger DO or  
counter generations.  
Connecting Digital I/O Signals on Non-Isolated Devices  
(NI 6124 Only) The DIO signals, P0.<0..7>, PFI <0..7>/P1.<0..7>, and  
PFI <8..15>/P2.<0..7> are referenced to D GND. You can individually  
program each line as an input or output. Figure 6-4 shows  
PFI <0..3>/P1.<0..3> configured for digital input and PFI <4..7>/P1.<4..7>  
configured for digital output. Digital input applications include receiving  
TTL signals and sensing external device states, such as the state of the  
switch shown in the figure. Digital output applications include sending  
TTL signals and driving external devices, such as the LED shown in the  
figure.  
© National Instruments Corporation  
6-9  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 6  
Digital I/O  
+5 V  
LED  
PFI <4..7>/  
P1.<4..7>  
PFI <0..3>/  
P1.<0..3>  
TTL Signal  
+5 V  
Switch  
D GND  
I/O Connector  
Non-Isolated S Series Device  
Figure 6-4. Digital I/O Connections  
Caution Exceeding the maximum input voltage ratings, which are listed in the  
specifications document for each non-isolated DAQ-STC2 S Series device, can damage the  
DAQ device and the computer. NI is not liable for any damage resulting from such signal  
connections.  
Getting Started with DIO Applications in Software on Non-Isolated  
Devices  
(NI 6124 Only) You can use non-isolated S Series devices in the following  
digital I/O applications:  
Static digital input  
Static digital output  
Digital waveform generation  
Digital waveform acquisition  
DI change detection  
Note For more information about programming digital I/O applications and triggers in  
software, refer to the NI-DAQmx Help or the LabVIEW Help in version 8.0 or later.  
NI 6124/6154 User Manual  
6-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 6  
Digital I/O  
Digital I/O for Isolated Devices  
(NI 6154 Only) S Series isolated devices contain ten lines of unidirectional  
DIO signals. The digital I/O port is comprised of six digital inputs and  
four digital outputs, all bank-isolated. Each digital line has the functionality  
of a PFI line. Input PFI lines can be used to input trigger signals to the  
different function modules of the DAQ-STC2 ASIC. The PFI pins also can  
be used as static digital inputs when not used to input triggers. Output PFI  
lines can export internal signals generated in any internal function module,  
as well as signals present in the RTSI bus. The PFI pins also can be used as  
static digital outputs when not used as trigger lines.  
The voltage input and output levels and the current drive levels of the DIO  
lines are listed in the NI 6154 Specifications.  
Figure 6-5 shows the circuitry of one bank-isolated DIO line.  
Isolation  
Barrier  
PFI/  
Static DI  
Digital  
Isolators  
PFI/  
Static DO  
Figure 6-5. Isolated S Series Devices Digital I/O Block Diagram  
Static DIO for Isolated Devices  
(NI 6154 Only) Isolated devices have unidirectional digital lines that are  
either static digital inputs (DI) or static digital outputs (DO). You can use  
DI and DO lines to monitor or control digital signals. All samples of static  
DI lines and updates of DO lines are software-timed. All DO lines are  
controlled by the same output enable. When a digital output line is enabled,  
all other digital output lines will also be enabled and driven to a default  
value of 0.  
You can select the up/down control input of general-purpose counters 0  
and 1 from any of the six digital input lines.  
© National Instruments Corporation  
6-11  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 6  
Digital I/O  
I/O Protection for Isolated Devices  
(NI 6154 Only) Each DIO and PFI signal is protected against over-voltage,  
under-voltage, and over-current conditions as well as ESD events.  
However, you should avoid these fault conditions by following these  
guidelines:  
Do not connect any DO line to any external signal source, ground  
signal, or power supply.  
Understand the current requirements of the load connected to DO  
signals. Do not exceed the specified current output limits of the DAQ  
device. NI has several signal conditioning solutions for digital  
applications requiring high-current drive.  
Do not drive any DI line with voltages outside of its normal operating  
range. The PFI or DIO lines have a smaller operating range than the AI  
signals.  
Treat the DAQ device as you would treat any static sensitive device.  
Always properly ground yourself and the equipment when handling  
the DAQ device or connecting to it.  
Connecting Digital I/O Signals on Isolated Devices  
(NI 6154 Only) The DIO signals, PFI <0..5>/P0.<0..5> and  
PFI <6..9>/P1.<0..3>, are referenced to D GND. PFI <0..5>/P0.<0..5> are  
inputs and PFI <6..9>/P1.<0..3> are outputs. Figure 6-6 shows digital  
inputs PFI <0..5>/P0.<0..5> and digital outputs PFI <6..9>/P1.<0..3>.  
Digital input applications include receiving TTL signals and sensing  
external device states, such as the state of the switch shown in the figure.  
Digital output applications include sending TTL signals and driving  
external devices, such as the LED shown in Figure 6-6.  
NI 6124/6154 User Manual  
6-12  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 6  
Digital I/O  
+5 V  
Isolation  
Barrier  
LED  
PFI <6..9>/P1.<0..3>  
Digital  
Isolators  
TTL Signal  
PFI <0..5>/P0.<0..5>  
+5 V  
Switch  
D GND  
I/O Connector  
Isolated S Series Device  
Figure 6-6. Isolated S Series Device Digital I/O Signal Connections  
Caution Exceeding the maximum input voltage ratings, which are listed in the NI 6154  
Specifications, can damage the DAQ device and the computer. NI is not liable for any  
damage resulting from such signal connections.  
Getting Started with DIO Applications in Software on Isolated Devices  
(NI 6154 Only) You can use isolated S Series devices in the following digital  
I/O applications:  
Static digital input  
Static digital output  
Note For more information about programming digital I/O applications and triggers in  
software, refer to the NI-DAQmx Help or the LabVIEW Help in version 8.0 or later.  
© National Instruments Corporation  
6-13  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
7
Counters  
S Series devices have two general-purpose 32-bit counter/timers and  
one frequency generator, as shown in Figure 7-1. The general-purpose  
counter/timers can be used for many measurement and pulse generation  
applications.  
Input Selection Muxes  
Counter 0  
Counter 0 Source (Counter 0 Timebase)  
Counter 0 Gate  
Counter 0 Internal Output  
Counter 0 Aux  
Counter 0 HW Arm  
Counter 0 A  
Counter 0 TC  
Counter 0 B (Counter 0 Up_Down)  
Counter 0 Z  
Input Selection Muxes  
Counter 1  
Counter 1 Source (Counter 1 Timebase)  
Counter 1 Gate  
Counter 0 Internal Output  
Counter 1 Aux  
Counter 1 HW Arm  
Counter 1 A  
Counter 0 TC  
Counter 1 B (Counter 1 Up_Down)  
Counter 1 Z  
Input Selection Muxes  
Frequency Generator  
Frequency Output Timebase  
Freq Out  
Figure 7-1. S Series Counters  
© National Instruments Corporation  
7-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
The counters have seven input signals, although in most applications only  
a few inputs are used.  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Counter Input Applications  
Counting Edges  
In edge counting applications, the counter counts edges on its Source after  
the counter is armed. You can configure the counter to count rising or  
falling edges on its Source input. You also can control the direction of  
counting (up or down).  
The counter values can be read on demand or with a sample clock.  
Single Point (On-Demand) Edge Counting  
With single point (on-demand) edge counting, the counter counts the  
number of edges on the Source input after the counter is armed. On-demand  
refers to the fact that software can read the counter contents at any time  
without disturbing the counting process. Figure 7-2 shows an example of  
single point edge counting.  
Counter Armed  
SOURCE  
Counter Value  
0
1
2
3
4
5
Figure 7-2. Single Point (On-Demand) Edge Counting  
You also can use a pause trigger to pause (or gate) the counter. When the  
pause trigger is active, the counter ignores edges on its Source input. When  
the pause trigger is inactive, the counter counts edges normally.  
You can route the pause trigger to the Gate input of the counter. You can  
configure the counter to pause counting when the pause trigger is high or  
when it is low. Figure 7-3 shows an example of on-demand edge counting  
with a pause trigger.  
NI 6124/6154 User Manual  
7-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 7  
Counters  
Counter Armed  
Pause Trigger  
(Pause When Low)  
SOURCE  
0
0
1
2
3
4
5
Counter Value  
Figure 7-3. Single Point (On-Demand) Edge Counting with Pause Trigger  
Buffered (Sample Clock) Edge Counting  
With buffered edge counting (edge counting using a sample clock), the  
counter counts the number of edges on the Source input after the counter is  
armed. The value of the counter is sampled on each active edge of a sample  
clock. A DMA controller transfers the sampled values to host memory.  
The count values returned are the cumulative counts since the counter  
armed event. That is, the sample clock does not reset the counter.  
You can route the counter sample clock to the Gate input of the counter. You  
can configure the counter to sample on the rising or falling edge of the  
sample clock.  
Figure 7-4 shows an example of buffered edge counting. Notice that  
counting begins when the counter is armed, which occurs before the first  
active edge on Gate.  
Counter Armed  
Sample Clock  
(Sample on Rising Edge)  
SOURCE  
Counter Value  
0
1
2
3
4
5
6
7
3
3
6
Buffer  
Figure 7-4. Buffered (Sample Clock) Edge Counting  
© National Instruments Corporation  
7-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 7  
Counters  
Controlling the Direction of Counting  
In edge counting applications, the counter can count up or down. You can  
configure the counter to do the following:  
Always count up  
Always count down  
Count up when the Counter n B input is high; count down when it is  
low  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Pulse-Width Measurement  
In pulse-width measurements, the counter measures the width of a pulse on  
its Gate input signal. You can configure the counter to measure the width  
of high pulses or low pulses on the Gate signal.  
You can route an internal or external periodic clock signal (with a known  
period) to the Source input of the counter. The counter counts the number  
of rising (or falling) edges on the Source signal while the pulse on the Gate  
signal is active.  
You can calculate the pulse width by multiplying the period of the Source  
signal by the number of edges returned by the counter.  
A pulse-width measurement will be accurate even if the counter is armed  
while a pulse train is in progress. If a counter is armed while the pulse is in  
the active state, it will wait for the next transition to the active state to begin  
the measurement.  
Single Pulse-Width Measurement  
With single pulse-width measurement, the counter counts the number of  
edges on the Source input while the Gate input remains active. When the  
Gate input goes inactive, the counter stores the count in a hardware save  
register and ignores other edges on the Gate and Source inputs. Software  
then reads the stored count.  
NI 6124/6154 User Manual  
7-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
Figure 7-5 shows an example of a single pulse-width measurement.  
GATE  
SOURCE  
Counter Value  
0
1
2
HW Save Register  
2
Figure 7-5. Single Pulse-Width Measurement  
Buffered Pulse-Width Measurement  
Buffered pulse-width measurement is similar to single pulse-width  
measurement, but buffered pulse-width measurement takes measurements  
over multiple pulses.  
The counter counts the number of edges on the Source input while the Gate  
input remains active. On each trailing edge of the Gate signal, the counter  
stores the count in a hardware save register. A DMA controller transfers the  
stored values to host memory.  
Figure 7-6 shows an example of a buffered pulse-width measurement.  
GATE  
SOURCE  
0
1
2
3
1
2
Counter Value  
3
3
2
Buffer  
3
2
Figure 7-6. Buffered Pulse-Width Measurement  
Note that if you are using an external signal as the Source, at least one  
Source pulse should occur between each active edge of the Gate signal.  
This condition ensures that correct values are returned by the counter. If this  
© National Instruments Corporation  
7-5  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 7  
Counters  
condition is not met, consider using duplicate count prevention, described  
in the Duplicate Count Prevention section.  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Period Measurement  
In period measurements, the counter measures a period on its Gate input  
signal after the counter is armed. You can configure the counter to measure  
the period between two rising edges or two falling edges of the Gate input  
signal.  
You can route an internal or external periodic clock signal (with a known  
period) to the Source input of the counter. The counter counts the number  
of rising (or falling) edges occurring on the Source input between the  
two active edges of the Gate signal.  
You can calculate the period of the Gate input by multiplying the period of  
the Source signal by the number of edges returned by the counter.  
Single Period Measurement  
With single period measurement, the counter counts the number of rising  
(or falling) edges on the Source input occurring between two active edges  
of the Gate input. On the second active edge of the Gate input, the counter  
stores the count in a hardware save register and ignores other edges on the  
Gate and Source inputs. Software then reads the stored count.  
Figure 7-7 shows an example of a single period measurement.  
GATE  
SOURCE  
Counter Value  
0
1
2
3
4
5
HW Save Register  
5
Figure 7-7. Single Period Measurement  
NI 6124/6154 User Manual  
7-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
Buffered Period Measurement  
Buffered period measurement is similar to single period measurement, but  
buffered period measurement measures multiple periods.  
The counter counts the number of rising (or falling) edges on the Source  
input between each pair of active edges on the Gate input. At the end of  
each period on the Gate signal, the counter stores the count in a hardware  
save register. A DMA controller transfers the stored values to host memory.  
The counter begins when it is armed. The arm usually occurs in the middle  
of a period of the Gate input. So the first value stored in the hardware save  
register does not reflect a full period of the Gate input. In most applications,  
this first point should be discarded.  
Figure 7-8 shows an example of a buffered period measurement.  
Counter Armed  
GATE  
SOURCE  
Counter Value  
1
2
1
2
3
1
2
3
1
2
3
3
2 (Discard)  
(Discard)  
2 (Discard)  
3
3
2
3
Buffer  
Note that if you are using an external signal as the Source, at least one  
Source pulse should occur between each active edge of the Gate signal.  
This condition ensures that correct values are returned by the counter. If this  
condition is not met, consider using duplicate count prevention, described  
in the Duplicate Count Prevention section.  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Semi-Period Measurement  
In semi-period measurements, the counter measures a semi-period on its  
Gate input signal after the counter is armed. A semi-period is the time  
between any two consecutive edges on the Gate input.  
© National Instruments Corporation  
7-7  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
You can route an internal or external periodic clock signal (with a known  
period) to the Source input of the counter. The counter counts the number  
of rising (or falling) edges occurring on the Source input between  
two edges of the Gate signal.  
You can calculate the semi-period of the Gate input by multiplying the  
period of the Source signal by the number of edges returned by the counter.  
Single Semi-Period Measurement  
Single semi-period measurement is equivalent to single pulse-width  
measurement.  
Buffered Semi-Period Measurement  
In buffered semi-period measurement, on each edge of the Gate signal, the  
counter stores the count in a hardware save register. A DMA controller  
transfers the stored values to host memory.  
The counter begins counting when it is armed. The arm usually occurs  
between edges on the Gate input. So the first value stored in the hardware  
save register does not reflect a full semi-period of the Gate input. In most  
applications, this first point should be discarded.  
Figure 7-9 shows an example of a buffered semi-period measurement.  
Counter Armed  
GATE  
SOURCE  
Counter Value  
0
1
2
1
2
3
1
1
2
1
2
3
2
2
2
3
1
2
2
3
1
2
3
Buffer  
1
Note that if you are using an external signal as the Source, at least one  
Source pulse should occur between each active edge of the Gate signal.  
This condition ensures that correct values are returned by the counter. If this  
condition is not met, consider using duplicate count prevention, described  
in the Duplicate Count Prevention section.  
NI 6124/6154 User Manual  
7-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Frequency Measurement  
You can use the counters to measure frequency in several different ways.  
You can choose one of the following methods depending on your  
application:  
Method 1: Measure Low Frequency with One Counter—In this  
method, you measure one period of your signal using a known  
timebase. This method is good for low frequency signals.  
You can route the signal to measure (F1) to the Gate of a counter. You  
can route a known timebase (Ft) to the Source of the counter. The  
known timebase can be 80MHzTimebase. For signals that might be  
slower than 0.02 Hz, use a slower known timebase.  
You can configure the counter to measure one period of the gate signal.  
The frequency of F1 is the inverse of the period. Figure 7-10 illustrates  
this method.  
Interval Measured  
F1  
F1  
Ft  
Gate  
1
2
3
N
Source  
Ft  
N
Single Period  
Measurement  
Period of F1 =  
Ft  
Ft  
Frequency of F1 =  
N
Figure 7-10. Method 1  
Method 1b: Measure Low Frequency with One Counter  
(Averaged)—In this method, you measure several periods of your  
signal using a known timebase. This method is good for low to medium  
frequency signals.  
You can route the signal to measure (F1) to the Gate of a counter. You  
can route a known timebase (Ft) to the Source of the counter. The  
known timebase can be 80MHzTimebase. For signals that might be  
slower than 0.02 Hz, use a slower known timebase.  
© National Instruments Corporation  
7-9  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
You can configure the counter to make K + 1 buffered period  
measurements. Recall that the first period measurement in the buffer  
should be discarded.  
Average the remaining K period measurements to determine the  
average period of F1. The frequency of F1 is the inverse of the average  
period. Figure 7-11 illustrates this method.  
Intervals Measured  
T1  
T2  
TK  
F1  
Ft  
Gate  
F1  
Ft  
12...N11......N2  
1......NK  
Source  
Buffered Period  
Measurement  
N1 + N2 + …NK  
1
×
Average Period of F1 =  
K
Ft  
K × Ft  
Frequency of F1 =  
N1 + N2 + …NK  
Figure 7-11. Method 1b  
Method 2: Measure High Frequency with Two Counters—In this  
method, you measure one pulse of a known width using your signal  
and derive the frequency of your signal from the result. This method is  
good for high frequency signals.  
In this method, you route a pulse of known duration (T) to the Gate of  
a counter. You can generate the pulse using a second counter. You also  
can generate the pulse externally and connect it to a PFI or RTSI  
terminal. You only need to use one counter if you generate the pulse  
externally.  
Route the signal to measure (F1) to the Source of the counter.  
Configure the counter for a single pulse-width measurement. If you  
measure the width of pulse T to be N periods of F1, the frequency of  
F1 is N/T.  
NI 6124/6154 User Manual  
7-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 7  
Counters  
Figure 7-12 illustrates this method. Another option would be to  
measure the width of a known period instead of a known pulse.  
Width of Pulse (T)  
Pulse  
Pulse  
Gate  
1
2
N
F1  
Source  
F1  
N
Pulse-Width  
Measurement  
Width of  
Pulse  
T =  
F1  
N
Frequency of F1 =  
T
Figure 7-12. Method 2  
Method 3: Measure Large Range of Frequencies Using  
Two Counters—By using two counters, you can accurately measure  
a signal that might be high or low frequency. This technique is called  
reciprocal frequency measurement. In this method, you generate a long  
pulse using the signal to measure. You then measure the long pulse  
with a known timebase. The S Series device can measure this long  
pulse more accurately than the faster input signal.  
© National Instruments Corporation  
7-11  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 7  
Counters  
You can route the signal to measure to the Source input of Counter 0,  
as shown in Figure 7-13. Assume this signal to measure has frequency  
F1. Configure Counter 0 to generate a single pulse that is the width of  
N periods of the source input signal.  
Signal to  
Measure (F1)  
SOURCE  
COUNTER 0  
OUT  
Signal of Known  
Frequency (F2)  
SOURCE  
OUT  
COUNTER 1  
GATE  
2
0
1
3
N  
CTR_0_SOURCE  
(Signal to Measure)  
CTR_0_OUT  
(CTR_1_GATE)  
Interval  
to Measure  
CTR_1_SOURCE  
Figure 7-13. Method 3  
Then route the Counter 0 Internal Output signal to the Gate input of  
Counter 1. You can route a signal of known frequency (F2) to the  
Counter 1 Source input. F2 can be 80MHzTimebase. For signals that  
might be slower than 0.02 Hz, use a slower known timebase. Configure  
Counter 1 to perform a single pulse-width measurement. Suppose the  
result is that the pulse width is J periods of the F2 clock.  
From Counter 0, the length of the pulse is N/F1. From Counter 1, the  
length of the same pulse is J/F2. Therefore, the frequency of F1 is  
given by F1 = F2 * (N/J).  
Choosing a Method for Measuring Frequency  
The best method to measure frequency depends on several factors  
including the expected frequency of the signal to measure, the desired  
accuracy, how many counters are available, and how long the measurement  
can take.  
NI 6124/6154 User Manual  
7-12  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 7  
Counters  
Method 1 uses only one counter. It is a good method for many  
applications. However, the accuracy of the measurement decreases as  
the frequency increases.  
Consider a frequency measurement on a 50 kHz signal using an  
80 MHz Timebase. This frequency corresponds to 1600 cycles of the  
80 MHz Timebase. Your measurement may return 1600 1 cycles  
depending on the phase of the signal with respect to the timebase. As  
your frequency becomes larger, this error of 1 cycle becomes more  
significant; Table 7-1 illustrates this point.  
Table 7-1. Frequency Measurement Method 1  
Task  
Equation  
F1  
Example 1  
50 kHz  
80 MHz  
1600  
Example 2  
5 MHz  
80 MHz  
16  
Actual Frequency to Measure  
Timebase Frequency  
Ft  
Actual Number of Timebase  
Periods  
Ft/F1  
Worst Case Measured Number  
of Timebase Periods  
(Ft/F1) – 1  
1599  
15  
Measured Frequency  
Error  
Ft F1/(Ft – F1)  
[Ft F1/(Ft – F1)] – F1  
[Ft/(Ft – F1)] – 1  
50.031 kHz  
31 Hz  
5.33 MHz  
333 kHz  
6.67%  
Error %  
0.06%  
Method 1b (measuring K periods of F1) improves the accuracy of the  
measurement. A disadvantage of Method 1b is that you have to take  
K + 1 measurements. These measurements take more time and  
consume some of the available PCI or PXI bandwidth.  
Method 2 is accurate for high frequency signals. However, the  
accuracy decreases as the frequency of the signal to measure  
decreases. At very low frequencies, Method 2 may be too inaccurate  
for your application. Another disadvantage of Method 2 is that it  
requires two counters (if you cannot provide an external signal of  
known width). An advantage of Method 2 is that the measurement  
completes in a known amount of time.  
Method 3 measures high and low frequency signals accurately.  
However, it requires two counters.  
© National Instruments Corporation  
7-13  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Chapter 7  
Counters  
Table 7-2 summarizes some of the differences in methods of measuring  
frequency.  
Table 7-2. Frequency Measurement Method Comparison  
Measures High  
Frequency  
Signals  
Measures Low  
Frequency  
Signals  
Number of  
Counters  
Used  
Number of  
Measurements  
Returned  
Method  
Accurately  
Accurately  
1
1b  
2
1
1
1
Many  
1
Poor  
Fair  
Good  
Good  
Poor  
1 or 2  
2
Good  
Good  
3
1
Good  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Position Measurement  
You can use the counters to perform position measurements with  
quadrature encoders or two-pulse encoders. You can measure angular  
position with X1, X2, and X4 angular encoders. Linear position can be  
measured with two-pulse encoders. You can choose to do either a single  
point (on-demand) position measurement or a buffered (sample clock)  
position measurement. You must arm a counter to begin position  
measurements.  
Measurements Using Quadrature Encoders  
The counters can perform measurements of quadrature encoders that use  
X1, X2, or X4 encoding. A quadrature encoder can have up to three  
channels—channels A, B, and Z.  
X1 Encoding—When channel A leads channel B in a quadrature  
cycle, the counter increments. When channel B leads channel A in a  
quadrature cycle, the counter decrements. The amount of increments  
and decrements per cycle depends on the type of encoding—X1, X2,  
or X4.  
NI 6124/6154 User Manual  
7-14  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 7  
Counters  
Figure 7-14 shows a quadrature cycle and the resulting increments and  
decrements for X1 encoding. When channel A leads channel B, the  
increment occurs on the rising edge of channel A. When channel B  
leads channel A, the decrement occurs on the falling edge of  
channel A.  
Ch A  
Ch B  
Counter Value  
6
7
7
6
5
Figure 7-14. X1 Encoding  
X2 Encoding—The same behavior holds for X2 encoding except the  
counter increments or decrements on each edge of channel A,  
depending on which channel leads the other. Each cycle results in  
two increments or decrements, as shown in Figure 7-15.  
Ch A  
Ch B  
Counter Value 5  
6
7
8
9
9
8
7
6
5
Figure 7-15. X2 Encoding  
X4 Encoding—Similarly, the counter increments or decrements on  
each edge of channels A and B for X4 encoding. Whether the counter  
increments or decrements depends on which channel leads the other.  
Each cycle results in four increments or decrements, as shown in  
Figure 7-16.  
Ch A  
Ch B  
Counter Value  
5
6
7
8
9
10 11 12 13 13 12 11 10  
9
8
7
6
5
Figure 7-16. X4 Encoding  
Channel Z Behavior  
Some quadrature encoders have a third channel, channel Z, which is also  
referred to as the index channel. A high level on channel Z causes the  
counter to be reloaded with a specified value in a specified phase of the  
quadrature cycle. You can program this reload to occur in any one of the  
four phases in a quadrature cycle.  
© National Instruments Corporation  
7-15  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 7  
Counters  
Channel Z behavior—when it goes high and how long it stays  
high—differs with quadrature encoder designs. You must refer to the  
documentation for your quadrature encoder to obtain timing of channel Z  
with respect to channels A and B. You must then ensure that channel Z is  
high during at least a portion of the phase you specify for reload. For  
instance, in Figure 7-17, channel Z is never high when channel A is high  
and channel B is low. Thus, the reload must occur in some other phase.  
In Figure 7-17, the reload phase is when both channel A and channel B are  
low. The reload occurs when this phase is true and channel Z is high.  
Incrementing and decrementing takes priority over reloading. Thus, when  
the channel B goes low to enter the reload phase, the increment occurs first.  
The reload occurs within one maximum timebase period after the reload  
phase becomes true. After the reload occurs, the counter continues to count  
as before. The figure illustrates channel Z reload with X4 decoding.  
Ch A  
Ch B  
Ch Z  
Max Timebase  
Counter Value  
5
6
7
8
9
0
1
2
3
4
A = 0  
B = 0  
Z = 1  
Figure 7-17. Channel Z Reload with X4 Decoding  
Measurements Using Two Pulse Encoders  
The counter supports two pulse encoders that have two channels—channels  
A and B.  
The counter increments on each rising edge of channel A. The counter  
decrements on each rising edge of channel B, as shown in Figure 7-18.  
Ch A  
Ch B  
Counter Value  
2
3
4
5
4
3
4
Figure 7-18. Measurements Using Two Pulse Encoders  
NI 6124/6154 User Manual  
7-16  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 7  
Counters  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Buffered (Sample Clock) Position Measurement  
With buffered position measurement (position measurement using a  
sample clock), the counter increments based on the encoding used after the  
counter is armed. The value of the counter is sampled on each active edge  
of a sample clock. A DMA controller transfers the sampled values to host  
memory. The count values returned are the cumulative counts since the  
counter armed event; that is, the sample clock does not reset the counter.  
You can route the counter sample clock to the Gate input of the counter.  
You can configure the counter to sample on the rising or falling edge of the  
sample clock.  
Figure 7-19 shows an example of a buffered edge X1 position  
measurement.  
Counter  
Armed  
Sample Clock  
(Sample on Rising Edge)  
Ch A  
Ch B  
Count  
Buffer  
0
1
2
4
5
1
1
4
Figure 7-19. Buffered Position Measurement  
Two-Signal Edge-Separation Measurement  
Two-signal edge-separation measurement is similar to pulse-width  
measurement, except that there are two measurement signals—Aux and  
Gate. An active edge on the Aux input starts the counting and an active  
edge on the Gate input stops the counting. You must arm a counter to begin  
a two edge separation measurement.  
After the counter has been armed and an active edge occurs on the Aux  
input, the counter counts the number of rising (or falling) edges on the  
Source. The counter ignores additional edges on the Aux input.  
The counter stops counting upon receiving an active edge on the Gate input.  
The counter stores the count in a hardware save register.  
© National Instruments Corporation  
7-17  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
You can configure the rising or falling edge of the Aux input to be the active  
edge. You can configure the rising or falling edge of the Gate input to be  
the active edge.  
Use this type of measurement to count events or measure the time that  
occurs between edges on two signals. This type of measurement is  
sometimes referred to as start/stop trigger measurement, second gate  
measurement, or A-to-B measurement.  
Single Two-Signal Edge-Separation Measurement  
With single two-signal edge-separation measurement, the counter counts  
the number of rising (or falling) edges on the Source input occurring  
between an active edge of the Gate signal and an active edge of the Aux  
signal. The counter then stores the count in a hardware save register and  
ignores other edges on its inputs. Software then reads the stored count.  
Figure 7-20 shows an example of a single two-signal edge-separation  
measurement.  
Counter  
Armed  
Measured Interval  
AUX  
GATE  
SOURCE  
Counter Value  
0
0
0
0
1
2
3
4
5
6
7
8
8
8
HW Save Register  
8
Figure 7-20. Single Two-Signal Edge-Separation Measurement  
Buffered Two-Signal Edge-Separation Measurement  
Buffered and single two-signal edge-separation measurements are similar,  
but buffered measurement measures multiple intervals.  
The counter counts the number of rising (or falling) edges on the Source  
input occurring between an active edge of the Gate signal and an active  
edge of the Aux signal. The counter then stores the count in a hardware save  
register. On the next active edge of the Gate signal, the counter begins  
another measurement. A DMA controller transfers the stored values to host  
memory.  
NI 6124/6154 User Manual  
7-18  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
Figure 7-21 shows an example of a buffered two-signal edge-separation  
measurement.  
AUX  
GATE  
SOURCE  
Counter Value  
1
2
3
1
2
3
1
2
3
3
3
3
3
3
3
Buffer  
Figure 7-21. Buffered Two-Signal Edge-Separation Measurement  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Counter Output Applications  
Simple Pulse Generation  
Single Pulse Generation  
The counter can output a single pulse. The pulse appears on the Counter n  
Internal Output signal of the counter.  
You can specify a delay from when the counter is armed to the beginning  
of the pulse. The delay is measured in terms of a number of active edges of  
the Source input.  
You can specify a pulse width. The pulse width is also measured in terms  
of a number of active edges of the Source input. You also can specify the  
active edge of the Source input (rising or falling).  
© National Instruments Corporation  
7-19  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 7  
Counters  
Figure 7-22 shows a generation of a pulse with a pulse delay of four and a  
pulse width of three (using the rising edge of Source).  
Counter Armed  
SOURCE  
OUT  
Figure 7-22. Single Pulse Generation  
Single Pulse Generation with Start Trigger  
The counter can output a single pulse in response to one pulse on a  
hardware Start Trigger signal. The pulse appears on the Counter n Internal  
Output signal of the counter.  
You can route the Start Trigger signal to the Gate input of the counter. You  
can specify a delay from the Start Trigger to the beginning of the pulse. You  
also can specify the pulse width. The delay and pulse width are measured  
in terms of a number of active edges of the Source input.  
After the Start Trigger signal pulses once, the counter ignores the Gate  
input.  
Figure 7-23 shows a generation of a pulse with a pulse delay of four and a  
pulse width of three (using the rising edge of Source).  
GATE  
(Start Trigger)  
SOURCE  
OUT  
Figure 7-23. Single Pulse Generation with Start Trigger  
Retriggerable Single Pulse Generation  
The counter can output a single pulse in response to each pulse on a  
hardware Start Trigger signal. The pulses appear on the Counter n Internal  
Output signal of the counter.  
NI 6124/6154 User Manual  
7-20  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 7  
Counters  
You can route the Start Trigger signal to the Gate input of the counter. You  
can specify a delay from the Start Trigger to the beginning of each pulse.  
You also can specify the pulse width. The delay and pulse width are  
measured in terms of a number of active edges of the Source input.  
The counter ignores the Gate input while a pulse generation is in progress.  
After the pulse generation is finished, the counter waits for another Start  
Trigger signal to begin another pulse generation.  
Figure 7-24 shows a generation of two pulses with a pulse delay of five and  
a pulse width of three (using the rising edge of Source).  
GATE  
(Start Trigger)  
SOURCE  
OUT  
Figure 7-24. Retriggerable Single Pulse Generation  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Pulse Train Generation  
Continuous Pulse Train Generation  
This function generates a train of pulses with programmable frequency and  
duty cycle. The pulses appear on the Counter n Internal Output signal of the  
counter.  
You can specify a delay from when the counter is armed to the beginning  
of the pulse train. The delay is measured in terms of a number of active  
edges of the Source input.  
You specify the high and low pulse widths of the output signal. The pulse  
widths are also measured in terms of a number of active edges of the Source  
input. You also can specify the active edge of the Source input (rising or  
falling).  
The counter can begin the pulse train generation as soon as the counter is  
armed, or in response to a hardware Start Trigger. You can route the Start  
Trigger to the Gate input of the counter.  
© National Instruments Corporation  
7-21  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
You also can use the Gate input of the counter as a Pause Trigger (if it is not  
used as a Start Trigger). The counter pauses pulse generation when the  
Pause Trigger is active.  
Figure 7-25 shows a continuous pulse train generation (using the rising  
edge of Source).  
SOURCE  
OUT  
Counter Armed  
Figure 7-25. Continuous Pulse Train Generation  
Continuous pulse train generation is sometimes called frequency division.  
If the high and low pulse widths of the output signal are M and N periods,  
then the frequency of the Counter n Internal Output signal is equal to the  
frequency of the Source input divided by M + N.  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Finite Pulse Train Generation  
This function generates a train of pulses of predetermined duration. This  
counter operation requires both counters. The first counter (for this  
example, Counter 0) generates a pulse of desired width. The second  
counter, Counter 1, generates the pulse train, which is gated by the pulse of  
the first counter. The routing is done internally. Figure 7-26 shows an  
example finite pulse train timing diagram.  
Counter 0  
(Paired Counter)  
Counter 1  
Generation  
Complete  
Figure 7-26. Finite Pulse Train Timing Diagram  
NI 6124/6154 User Manual  
7-22  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 7  
Counters  
Frequency Generation  
You can generate a frequency by using a counter in pulse train generation  
mode or by using the frequency generator circuit.  
The frequency generator can output a square wave at many different  
frequencies. The frequency generator is independent of the  
two general-purpose 32-bit counter/timer modules on S Series devices.  
Figure 7-27 shows a block diagram of the frequency generator.  
Frequency  
Output  
Timebase  
20 MHz Timebase  
÷ 2  
Frequency Generator  
FREQ OUT  
100 kHz Timebase  
Divisor  
(1–16)  
Figure 7-27. Frequency Generator Block Diagram  
The frequency generator generates the Frequency Output signal. The  
Frequency Output signal is the Frequency Output Timebase divided by a  
number you select from 1 to 16. The Frequency Output Timebase can be  
either the 20 MHz Timebase divided by 2 or the 100 kHz Timebase.  
The duty cycle of Frequency Output is 50% if the divider is either 1 or  
an even number. For an odd divider, suppose the divider is set to D.  
In this case, Frequency Output is low for (D + 1)/2 cycles and high for  
(D – 1)/2 cycles of the Frequency Output Timebase.  
Figure 7-28 shows the output waveform of the frequency generator when  
the divider is set to 5.  
Frequency  
Output  
Timebase  
FREQ OUT  
(Divisor = 5)  
Figure 7-28. Frequency Generator Output Waveform  
© National Instruments Corporation  
7-23  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 7  
Counters  
Frequency Output can be routed out to any PFI <0..15> or RTSI <0..7>  
terminal. All PFI terminals are set to high-impedance at startup. The FREQ  
OUT signal also can be routed to DO Sample Clock and DI Sample Clock.  
In software, program the frequency generator as you would program one of  
the counters for pulse train generation.  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Frequency Division  
The counters can generate a signal with a frequency that is a fraction of  
an input signal. This function is equivalent to continuous pulse train  
generation. Refer to the Continuous Pulse Train Generation section for  
detailed information.  
For information about connecting counter signals, refer to the Default  
Counter/Timer Pinouts section.  
Pulse Generation for ETS  
In the equivalent time sampling (ETS) application, the counter produces a  
pulse on the output a specified delay after an active edge on Gate. After  
each active edge on Gate, the counter cumulatively increments the delay  
between the Gate and the pulse on the output by a specified amount. Thus,  
the delay between the Gate and the pulse produced successively increases.  
The increase in the delay value can be between 0 and 255. For instance, if  
you specify the increment to be 10, the delay between the active Gate edge  
and the pulse on the output will increase by 10 every time a new pulse is  
generated.  
Suppose you program your counter to generate pulses with a delay of 100  
and pulse width of 200 each time it receives a trigger. Furthermore, suppose  
you specify the delay increment to be 10. On the first trigger, your pulse  
delay will be 100, on the second it will be 110, on the third it will be 120;  
the process will repeat in this manner until the counter is disarmed. The  
counter ignores any Gate edge that is received while the pulse triggered by  
the previous Gate edge is in progress.  
NI 6124/6154 User Manual  
7-24  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 7  
Counters  
The waveform thus produced at the counter’s output can be used to provide  
timing for undersampling applications where a digitizing system can  
sample repetitive waveforms that are higher in frequency than the Nyquist  
frequency of the system. Figure 7-29 shows an example of pulse generation  
for ETS; the delay from the trigger to the pulse increases after each  
subsequent Gate active edge.  
GATE  
OUT  
D1  
D2 = D1 + ΔD  
D3 = D1 + 2ΔD  
Figure 7-29. Pulse Generation for ETS  
For information about connecting counter signals, refer to the Default  
Counter Timing Signals  
S Series devices feature the following counter timing signals:  
Counter n Source Signal  
Counter n Gate Signal  
Counter n Aux Signal  
Counter n A Signal  
Counter n B Signal  
Counter n Z Signal  
Counter n Up_Down Signal  
Counter n HW Arm Signal  
Counter n Internal Output Signal  
Counter n TC Signal  
Frequency Output Signal  
In this section, n refers to either Counter 0 or 1. For example, Counter n  
Source refers to two signals—Counter 0 Source (the source input to  
Counter 0) and Counter 1 Source (the source input to Counter 1).  
© National Instruments Corporation  
7-25  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 7  
Counters  
Counter n Source Signal  
The selected edge of the Counter n Source signal increments and  
decrements the counter value depending on the application the counter  
is performing. Table 7-3 lists how this terminal is used in various  
applications.  
Table 7-3. Counter Applications and Counter n Source  
Application  
Purpose of Source Terminal  
Counter Timebase  
Counter Timebase  
Input Terminal  
Pulse Generation  
One Counter Time Measurements  
Two Counter Time Measurements  
Non-Buffered Edge Counting  
Buffered Edge Counting  
Two-Edge Separation  
Input Terminal  
Input Terminal  
Counter Timebase  
Routing a Signal to Counter n Source  
Each counter has independent input selectors for the Counter n Source  
signal. Any of the following signals can be routed to the Counter n Source  
input:  
80 MHz Timebase  
20 MHz Timebase  
100 kHz Timebase  
RTSI <0..7>  
PFI <0..15>  
PXI_CLK10  
PXI_STAR  
Analog Comparison Event  
In addition, Counter 1 TC or Counter 1 Gate can be routed to  
Counter 0 Source. Counter 0 TC or Counter 0 Gate can be routed to  
Counter 1 Source.  
Some of these options may not be available in some driver software.  
NI 6124/6154 User Manual  
7-26  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 7  
Counters  
Routing Counter n Source to an Output Terminal  
You can route Counter n Source out to any PFI <0..15> or RTSI <0..7>  
terminal. All PFIs are set to high-impedance at startup.  
Counter n Gate Signal  
The Counter n Gate signal can perform many different operations  
depending on the application including starting and stopping the counter,  
and saving the counter contents.  
Routing a Signal to Counter n Gate  
Each counter has independent input selectors for the Counter n Gate signal.  
Any of the following signals can be routed to the Counter n Gate input:  
RTSI <0..7>  
PFI <0..15>  
AI Reference Trigger (ai/ReferenceTrigger)  
AI Start Trigger (ai/StartTrigger)  
AI Sample Clock (ai/SampleClock)  
AI Convert Clock (ai/ConvertClock)  
AO Sample Clock (ao/SampleClock)  
DI Sample Clock (di/SampleClock)  
DO Sample Clock (do/SampleClock)  
PXI_STAR  
Change Detection Event  
Analog Comparison Event  
In addition, Counter 1 Internal Output or Counter 1 Source can be routed to  
Counter 0 Gate. Counter 0 Internal Output or Counter 0 Source can be  
routed to Counter 1 Gate.  
Some of these options may not be available in some driver software.  
Routing Counter n Gate to an Output Terminal  
You can route Counter n Gate out to any PFI <0..15> or RTSI <0..7>  
terminal. All PFIs are set to high-impedance at startup.  
© National Instruments Corporation  
7-27  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
Counter n Aux Signal  
The Counter n Aux signal indicates the first edge in a two-signal  
edge-separation measurement.  
Routing a Signal to Counter n Aux  
Each counter has independent input selectors for the Counter n Aux signal.  
Any of the following signals can be routed to the Counter n Aux input:  
RTSI <0..7>  
PFI <0..15>  
AI Reference Trigger (ai/ReferenceTrigger)  
AI Start Trigger (ai/StartTrigger)  
PXI_STAR  
Analog Comparison Event  
In addition, Counter 1 Internal Output, Counter 1 Gate, Counter 1 Source,  
or Counter 0 Gate can be routed to Counter 0 Aux. Counter 0 Internal  
Output, Counter 0 Gate, Counter 0 Source, or Counter 1 Gate can be routed  
to Counter 1 Aux.  
Some of these options may not be available in some driver software.  
Counter n A, Counter n B, and Counter n Z Signals  
Counter n B can control the direction of counting in edge counting  
applications. Use the A, B, and Z inputs to each counter when measuring  
quadrature encoders or measuring two pulse encoders.  
Routing Signals to A, B, and Z Counter Inputs  
Each counter has independent input selectors for each of the A, B, and  
Z inputs. Any of the following signals can be routed to each input:  
RTSI <0..7>  
PFI <0..15>  
PXI_STAR  
Analog Comparison Event  
Routing Counter n Z Signal to an Output Terminal  
You can route Counter n Z out to RTSI <0..7>.  
NI 6124/6154 User Manual  
7-28  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 7  
Counters  
Counter n Up_Down Signal  
Counter n Up_Down is another name for the Counter n B signal.  
Counter n HW Arm Signal  
The Counter n HW Arm signal enables a counter to begin an input or output  
function.  
To begin any counter input or output function, you must first enable, or arm,  
the counter. In some applications, such as buffered semi-period  
measurement, the counter begins counting when it is armed. In other  
applications, such as single pulse-width measurement, the counter begins  
waiting for the Gate signal when it is armed. Counter output operations can  
use the arm signal in addition to a start trigger.  
Software can arm a counter or configure counters to be armed on a  
hardware signal. Software calls this hardware signal the Arm Start Trigger.  
Internally, software routes the Arm Start Trigger to the Counter n HW Arm  
input of the counter.  
Routing Signals to Counter n HW Arm Input  
Any of the following signals can be routed to the Counter n HW Arm input:  
RTSI <0..7>  
PFI <0..15>  
AI Reference Trigger (ai/ReferenceTrigger)  
AI Start Trigger (ai/StartTrigger)  
PXI_STAR  
Analog Comparison Event  
Counter 1 Internal Output can be routed to Counter 0 HW Arm. Counter 0  
Internal Output can be routed to Counter 1 HW Arm.  
Some of these options may not be available in some driver software.  
Counter n Internal Output and Counter n TC Signals  
The Counter n Internal Output signal changes in response to Counter n TC.  
The two software-selectable output options are pulse output on TC and  
toggle output on TC. The output polarity is software-selectable for both  
options.  
© National Instruments Corporation  
7-29  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 7  
Counters  
With pulse or pulse train generation tasks, the counter drives the pulse(s) on  
the Counter n Internal Output signal. The Counter n Internal Output signal  
can be internally routed to be a counter/timer input or an “external” source  
for AI, AO, DI, or DO timing signals.  
Routing Counter n Internal Output to an Output  
Terminal  
You can route Counter n Internal Output to any PFI <0..15> or  
RTSI <0..7> terminal. All PFIs are set to high-impedance at startup.  
Frequency Output Signal  
The Frequency Output (FREQ OUT) signal is the output of the frequency  
output generator.  
Routing Frequency Output to a Terminal  
You can route Frequency Output to any PFI <0..15> terminal. All PFIs are  
set to high-impedance at startup. The FREQ OUT signal also can be routed  
to DO Sample Clock and DI Sample Clock.  
Default Counter/Timer Pinouts  
By default, NI-DAQmx routes the counter/timer inputs and outputs to the  
PFI pins. Refer to Appendix A, Device-Specific Information, for the default  
counter/timer pin table for your device.  
You can use these defaults or select other sources and destinations for the  
counter/timer signals in NI-DAQmx. Refer to Connecting Counter Signals  
in the NI-DAQmx Help or the LabVIEW Help in version 8.0 or later for  
more information about how to connect your signals for common counter  
measurements and generations. DAQ-STC2-based S Series default PFI  
lines for counter functions are listed in Physical Channels in the  
NI-DAQmx Help or the LabVIEW Help in version 8.0 or later.  
NI 6124/6154 User Manual  
7-30  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 7  
Counters  
Counter Triggering  
Counters support three different triggering actions:  
Arm Start Trigger—To begin any counter input or output function,  
you must first enable, or arm, the counter. Software can arm a counter  
or configure counters to be armed on a hardware signal. Software calls  
this hardware signal the Arm Start Trigger. Internally, software routes  
the Arm Start Trigger to the Counter n HW Arm input of the counter.  
For counter output operations, you can use it in addition to the start and  
pause triggers. For counter input operations, you can use the arm start  
trigger to have start trigger-like behavior. The arm start trigger can be  
used for synchronizing multiple counter input and output tasks.  
When using an arm start trigger, the arm start trigger source is routed  
to the Counter n HW Arm signal.  
Start Trigger—For counter output operations, a start trigger can be  
configured to begin a finite or continuous pulse generation. Once a  
continuous generation has triggered, the pulses continue to generate  
until you stop the operation in software. For finite generations, the  
specified number of pulses is generated and the generation stops unless  
you use the retriggerable attribute. When you use this attribute,  
subsequent start triggers cause the generation to restart.  
When using a start trigger, the start trigger source is routed to the  
Counter n Gate signal input of the counter.  
Counter input operations can use the arm start trigger to have start  
trigger-like behavior.  
Pause TriggerYou can use pause triggers in edge counting and  
continuous pulse generation applications. For edge counting  
acquisitions, the counter stops counting edges while the external  
trigger signal is low and resumes when the signal goes high or vice  
versa. For continuous pulse generations, the counter stops generating  
pulses while the external trigger signal is low and resumes when the  
signal goes high or vice versa.  
When using a pause trigger, the pause trigger source is routed to the  
Counter n Gate signal input of the counter.  
© National Instruments Corporation  
7-31  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
Other Counter Features  
Cascading Counters  
You can internally route the Counter n Internal Output and Counter n TC  
signals of each counter to the Gate inputs of the other counter. By cascading  
two counters together, you can effectively create a 64-bit counter. By  
cascading counters, you also can enable other applications. For example, to  
improve the accuracy of frequency measurements, use reciprocal  
frequency measurement, as described in the Method 3 bullet in the  
Frequency Measurement section.  
Counter Filters  
You can enable a programmable debouncing filter on each PFI, RTSI, or  
PXI_STAR signal. When the filters are enabled, your device samples the  
input on each rising edge of a filter clock. S Series devices use an onboard  
oscillator to generate the filter clock with a 40 MHz frequency.  
Note NI-DAQmx only supports filters on counter inputs.  
The following is an example of low to high transitions of the input signal.  
High to low transitions work similarly.  
Assume that an input terminal has been low for a long time. The input  
terminal then changes from low to high, but glitches several times. When  
the filter clock has sampled the signal high on N consecutive edges, the low  
to high transition is propagated to the rest of the circuit. The value of N  
depends on the filter setting; refer to Table 7-4.  
Table 7-4. Filters  
N (Filter Clocks  
Needed to  
Pass Signal)  
Pulse Width  
Guaranteed to  
Pass Filter  
Pulse Width  
Guaranteed to  
Not Pass Filter  
Filter Setting  
125 ns  
5
257  
125 ns  
6.425 μs  
2.56 ms  
100 ns  
6.400 μs  
2.54 ms  
6.425 μs  
2.56 ms  
~101,800  
NI 6124/6154 User Manual  
7-32  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 7  
Counters  
The filter setting for each input can be configured independently. On power  
up, the filters are disabled. Figure 7-30 shows an example of a low to high  
transition on an input that has its filter set to 125 ns (N = 5).  
RTSI, PFI, or  
PXI_STAR Terminal  
Filtered input goes  
high when terminal  
is sampled high on  
five consecutive filter  
clocks.  
1
1
2
3
4
1
2
3
4
5
Filter Clock  
(40 MHz)  
Filtered Input  
Figure 7-30. Filter Example  
Enabling filters introduces jitter on the input signal. For the 125 ns and  
6.425 μs filter settings, the jitter is up to 25 ns. On the 2.56 ms setting, the  
jitter is up to 10.025 μs.  
When a PFI input is routed directly to RTSI, or a RTSI input is routed  
directly to PFI, the S Series device does not use the filtered version of the  
input signal.  
Refer to the KnowledgeBase document, Digital Filtering with M Series  
and CompactDAQ, for more information about digital filters and counters.  
To access this KnowledgeBase, go to ni.com/infoand enter the info  
code rddfms.  
Prescaling  
Prescaling allows the counter to count a signal that is faster than the  
maximum timebase of the counter. S Series devices offer 8X and 2X  
prescaling on each counter (prescaling can be disabled). Each prescaler  
consists of a small, simple counter that counts to eight (or two) and rolls  
over. This counter can run faster than the larger counters, which simply  
count the rollovers of this smaller counter. Thus, the prescaler acts as a  
frequency divider on the Source and puts out a frequency that is one-eighth  
(or one-half) of what it is accepting.  
© National Instruments Corporation  
7-33  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 7  
Counters  
External Signal  
Prescaler Rollover  
(Used as Source  
by Counter)  
Counter Value  
0
1
Figure 7-31. Prescaling  
Prescaling is intended to be used for frequency measurement where the  
measurement is made on a continuous, repetitive signal. The prescaling  
counter cannot be read; therefore, you cannot determine how many edges  
have occurred since the previous rollover. Prescaling can be used for event  
counting provided it is acceptable to have an error of up to seven (or one).  
Prescaling can be used when the counter Source is an external signal.  
Prescaling is not available if the counter Source is one of the internal  
timebases (80MHzTimebase, 20MHzTimebase, or 100kHzTimebase).  
Duplicate Count Prevention  
Duplicate count prevention (or synchronous counting mode) ensures that a  
counter returns correct data in applications that use a slow or non-periodic  
external source. Duplicate count prevention applies only to buffered  
counter applications such as measuring frequency or period. In such  
buffered applications, the counter should store the number of times an  
external Source pulses between rising edges on the Gate signal.  
NI 6124/6154 User Manual  
7-34  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 7  
Counters  
Example Application That Works Correctly  
(No Duplicate Counting)  
Figure 7-32 shows an external buffered signal as the period measurement  
Source.  
Rising Edge  
of Gate  
Counter detects rising  
edge of Gate on the next  
rising edge of Source.  
Gate  
Source  
Counter Value  
6
7
1
2
1
7
2
7
Buffer  
Figure 7-32. Duplicate Count Prevention Example  
On the first rising edge of the Gate, the current count of 7 is stored. On the  
next rising edge of the Gate, the counter stores a 2 since two Source pulses  
occurred after the previous rising edge of Gate.  
The counter synchronizes or samples the Gate signal with the Source  
signal, so the counter does not detect a rising edge in the Gate until the next  
Source pulse. In this example, the counter stores the values in the buffer on  
the first rising Source edge after the rising edge of Gate. The details of  
when exactly the counter synchronizes the Gate signal vary depending on  
the synchronization mode. Synchronization modes are described in the  
Synchronization Modes section.  
© National Instruments Corporation  
7-35  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 7  
Counters  
Example Application That Works Incorrectly  
(Duplicate Counting)  
In Figure 7-33, after the first rising edge of Gate, no Source pulses occur,  
so the counter does not write the correct data to the buffer.  
No Source edge, so no  
value written to buffer.  
Gate  
Source  
6
7
1
Counter Value  
7
Buffer  
Figure 7-33. Duplicate Count Example  
Example Application That Prevents Duplicate Count  
With duplicate count prevention enabled, the counter synchronizes both the  
Source and Gate signals to the 80 MHz Timebase. By synchronizing to the  
timebase, the counter detects edges on the Gate even if the Source does not  
pulse. This enables the correct current count to be stored in the buffer even  
if no Source edges occur between Gate signals, as shown in Figure 7-34.  
Counter detects  
rising Gate edge.  
Counter value  
increments only  
one time for each  
Source pulse.  
Gate  
Source  
80 MHz Timebase  
Counter Value  
Buffer  
6
7
0
1
7
0
7
Figure 7-34. Duplicate Count Prevention Example  
NI 6124/6154 User Manual  
7-36  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
Even if the Source pulses are long, the counter increments only once for  
each Source pulse.  
Normally, the counter value and Counter n Internal Output signals change  
synchronously to the Source signal. With duplicate count prevention, the  
counter value and Counter n Internal Output signals change synchronously  
to the 80 MHz Timebase.  
Note that duplicate count prevention should only be used if the frequency  
of the Source signal is 20 MHz or less.  
When To Use Duplicate Count Prevention  
You should use duplicate count prevention if the following conditions are  
true:  
You are making a counter measurement.  
You are using an external signal (such as PFI x) as the counter Source.  
The frequency of the external source is 20 MHz or less.  
You can have the counter value and output to change synchronously  
with the 80 MHz Timebase.  
In all other cases, you should not use duplicate count prevention.  
Enabling Duplicate Count Prevention in NI-DAQmx  
You can enable duplicate count prevention in NI-DAQmx by setting the  
Enable Duplicate Count Prevention attribute/property. For specific  
information about finding the Enable Duplicate Count Prevention  
attribute/property, refer to the help file for the API you are using.  
Synchronization Modes  
The 32-bit counter counts up or down synchronously with the Source  
signal. The Gate signal and other counter inputs are asynchronous to the  
Source signal. So S Series devices synchronize these signals before  
presenting them to the internal counter.  
S Series devices use one of three synchronization methods:  
80 MHz source mode  
Other internal source mode  
External source mode  
© National Instruments Corporation  
7-37  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 7  
Counters  
In DAQmx, the device uses 80 MHz source mode if you perform the  
following:  
Perform a position measurement  
Select duplicate count prevention  
Otherwise, the mode depends on the signal that drives Counter n Source.  
Table 7-5 describes the conditions for each mode.  
Table 7-5. Synchronization Mode Conditions  
Duplicate Count  
Prevention Enabled  
Type of  
Measurement  
Signal Driving  
Counter n Source  
Synchronization  
Mode  
Yes  
No  
No  
No  
Any  
Any  
80 MHz Source  
80 MHz Source  
80 MHz Source  
Other Internal Source  
Position Measurement Any  
Any  
80 MHz Timebase  
All Except Position  
Measurement  
20 MHz Timebase,  
100 kHz Timebase,  
or PXI_CLK10  
No  
All Except Position  
Measurement  
Any Other Signal  
(such as PFI or RTSI)  
External Source  
80 MHz Source Mode  
In 80 MHz source mode, the device synchronizes signals on the rising edge  
of the source, and counts on the following rising edge of the source,  
as shown in Figure 7-35.  
Source  
Synchronize  
Figure 7-35. 80 MHz Source Mode  
Count  
NI 6124/6154 User Manual  
7-38  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 7  
Counters  
Other Internal Source Mode  
In other internal source mode, the device synchronizes signals on the  
falling edge of the source, and counts on the following rising edge of the  
source, as shown in Figure 7-36.  
Source  
Synchronize  
Count  
Figure 7-36. Other Internal Source Mode  
External Source Mode  
In external source mode, the device generates a delayed Source signal by  
delaying the Source signal by several nanoseconds. The device  
synchronizes signals on the rising edge of the delayed Source signal, and  
counts on the following rising edge of the source, as shown in Figure 7-37.  
Source  
Synchronize  
Delayed Source  
Count  
Figure 7-37. External Source Mode  
© National Instruments Corporation  
7-39  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
8
Programmable Function  
Interfaces (PFI)  
Refer to one of the following sections, depending on your device:  
PFI for Non-Isolated Devices—NI 6124 devices have 16 PFI pins in  
addition to eight lines of bidirectional DIO signals.  
PFI for Isolated Devices—NI 6154 devices have 10 equivalent  
directional PFI pins that can be independently configured as an input  
or output.  
PFI for Non-Isolated Devices  
(NI 6124 Only) Non-isolated S Series devices have 16 Programmable  
Function Interface (PFI) signals. In addition, these devices have eight lines  
of bidirectional DIO signals.  
Each PFI can be individually configured as the following:  
A static digital input  
A static digital output  
A timing input signal for AI, AO, DI, DO, or counter/timer functions  
A timing output signal from AI, AO, DI, DO, or counter/timer  
functions  
© National Instruments Corporation  
8-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 8  
Programmable Function Interfaces (PFI)  
Each PFI input also has a programmable debouncing filter. Figure 8-1  
shows the circuitry of one PFI line. Each PFI line is similar.  
Timing Signals  
Static DO  
Buffer  
PFI x/P1 or  
PFI x/P2  
I/O Protection  
Direction Control  
Static DI  
Weak Pull-Down  
To Input Timing  
Signal Selectors  
PFI  
Filters  
Figure 8-1. PFI Circuitry on Non-Isolated S Series Devices  
When a terminal is used as a timing input or output signal, it is called PFI x  
(where x is an integer from 0 to 15). When a terminal is used as a static  
digital input or output, it is called P1.x or P2.x. On the I/O connector, each  
terminal is labeled PFI x/P1.x or PFI x/P2.x.  
The voltage input and output levels and the current drive levels of the PFI  
signals are listed in the NI 6124 Specifications.  
PFI for Isolated Devices  
(NI 6154 Only) Isolated S Series devices have 10 Programmable Function  
Interface (PFI) signals—six input signals and four output signals.  
Each PFI <0..5>/P0.<0..5> can be configured as a timing input signal for  
AI or counter/timer functions or a static digital input. Each PFI input also  
has a programmable debouncing filter.  
NI 6124/6154 User Manual  
8-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 8  
Programmable Function Interfaces (PFI)  
Figure 8-2 shows the circuitry of one PFI input line. Each PFI line is  
similar.  
Isolation  
Barrier  
Static DI  
Digital  
Isolators  
PFI <0..5>/P0.<0..5>  
I/O Protection  
To Input Timing  
Signal Selectors  
PFI  
Filters  
Figure 8-2. PFI Input Circuitry on Isolated S Series Devices  
Each PFI <6..9>/P1.<0..3> can be configured as a timing output signal  
from AI or counter/timer functions or a static digital output.  
Figure 8-3 shows the circuitry of one PFI output line. Each PFI line is  
similar.  
Isolation  
Barrier  
Timing Signals  
Digital  
PFI <6..9>/P1.<0..3>  
I/O Protection  
Isolators  
Static DO  
Buffer  
Output  
Enable  
Note: One output enable is shared  
by all digital output signals.  
Figure 8-3. PFI Output Circuitry on Isolated S Series Devices  
When a terminal is used as a timing input or output signal, it is called PFI x  
(where x is an integer from 0 to 9). When a terminal is used as a static digital  
input or output, it is called P0.x or P1.x.  
The voltage input and output levels and the current drive levels of the PFI  
signals are listed in the NI 6154 Specifications.  
© National Instruments Corporation  
8-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 8  
Programmable Function Interfaces (PFI)  
Using PFI Terminals as Timing Input Signals  
Use PFI terminals to route external timing signals to many different  
S Series functions.1 Each PFI terminal (or input PFI terminal) can be routed  
to any of the following signals:  
AI Convert Clock (ai/ConvertClock)  
AI Sample Clock (ai/SampleClock)  
AI Start Trigger (ai/StartTrigger)  
AI Reference Trigger (ai/ReferenceTrigger)  
AI Sample Clock Timebase (ai/SampleClockTimebase)  
AO Start Trigger (ao/StartTrigger)  
AO Sample Clock (ao/SampleClock)  
AO Sample Clock Timebase (ao/SampleClockTimebase)  
AO Pause Trigger (ao/PauseTrigger)  
Counter input signals for either counter—Source, Gate, Aux,  
HW_Arm, A, B, Z  
(NI 6124 Only) DI Sample Clock (di/SampleClock)  
(NI 6124 Only) DO Sample Clock (do/SampleClock)  
Most functions allow you to configure the polarity of PFI inputs and  
whether the input is edge or level sensitive.  
Exporting Timing Output Signals Using PFI Terminals  
You can route any of the following timing signals to PFI terminals  
configured as an output:2  
AI Convert Clock* (ai/ConvertClock)  
AI Hold Complete Event (ai/HoldCompleteEvent)  
AI Reference Trigger (ai/ReferenceTrigger)  
AI Sample Clock (ai/SampleClock)  
AI Start Trigger (ai/StartTrigger)  
AO Sample Clock* (ao/SampleClock)  
1
2
You can use any PFI terminal as a timing input signal on NI 6124 devices. You can use PFI <0..5> terminals as timing input  
signals on NI 6154 devices.  
You can export timing output signals with any PFI terminal on NI 6124 devices. You can export timing output signals with  
PFI <6..9> on NI 6154 devices.  
NI 6124/6154 User Manual  
8-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 8  
Programmable Function Interfaces (PFI)  
AO Start Trigger (ao/StartTrigger)  
Counter n Source  
Counter n Gate  
Counter n Internal Output  
Frequency Output  
PXI_STAR  
RTSI <0..7>  
(NI 6124 Only) Analog Comparison Event  
(NI 6124 Only) Change Detection Event  
(NI 6124 Only) DI Sample Clock* (di/SampleClock)  
(NI 6124 Only) DO Sample Clock* (do/SampleClock)  
Note Signals with a * are inverted before being driven to a terminal; that is, these signals  
are active low.  
Using PFI Terminals as Static Digital Inputs and Outputs  
You can configure PFI terminals to be static digital input or output  
terminals:  
NI 6124 Devices—Each PFI can be individually configured as a static  
digital input or a static digital output. When a terminal is used as a  
static digital input or output, it is called P1.x or P2.x. On the I/O  
connector, each terminal is labeled PFI x/P1.x or PFI x/P2.x.  
In addition, NI 6124 devices have eight lines of bidirectional DIO  
signals.  
NI 6154 Devices— When a terminal is used as a static digital input or  
output, it is called P0.x or P1.x. On the I/O connector, each terminal is  
labeled PFI x/P0.x or PFI x/P1.x.  
© National Instruments Corporation  
8-5  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 8  
Programmable Function Interfaces (PFI)  
Connecting PFI Input Signals  
All PFI input connections are referenced to D GND. Figure 8-4 shows this  
reference, and how to connect an external PFI 0 source and an external  
PFI 2 source to two PFI terminals.  
PFI 0  
PFI 2  
PFI 0  
PFI 2  
Source  
Source  
D GND  
S Series Device  
I/O Connector  
Figure 8-4. PFI Input Signals Connections  
PFI Filters  
You can enable a programmable debouncing filter on each PFI, RTSI, or  
PXI_STAR signal. When the filters are enabled, your device samples the  
input on each rising edge of a filter clock. These S Series devices use an  
onboard oscillator to generate the filter clock with a 40 MHz frequency.  
Note NI-DAQmx only supports filters on counter inputs.  
The following is an example of low to high transitions of the input signal.  
High to low transitions work similarly.  
NI 6124/6154 User Manual  
8-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 8  
Programmable Function Interfaces (PFI)  
Assume that an input terminal has been low for a long time. The input  
terminal then changes from low to high, but glitches several times. When  
the filter clock has sampled the signal high on N consecutive edges, the low  
to high transition is propagated to the rest of the circuit. The value of N  
depends on the filter setting; refer to Table 8-1.  
Table 8-1. Filters  
N (Filter Clocks  
Needed to  
Pass Signal)  
Pulse Width  
Guaranteed to  
Pass Filter  
Pulse Width  
Guaranteed to  
Not Pass Filter  
Filter Setting  
125 ns  
5
257  
125 ns  
6.425 μs  
2.56 ms  
100 ns  
6.400 μs  
2.54 ms  
6.425 μs  
2.56 ms  
~101,800  
Disabled  
The filter setting for each input can be configured independently. On power  
up, the filters are disabled. Figure 8-5 shows an example of a low to high  
transition on an input that has its filter set to 125 ns (N = 5).  
RTSI, PFI, or  
PXI_STAR Terminal  
Filtered input goes  
high when terminal  
is sampled high on  
five consecutive filter  
clocks.  
1
1
2
3
4
1
2
3
4
5
Filter Clock  
(40 MHz)  
Filtered Input  
Figure 8-5. Filter Example  
Enabling filters introduces jitter on the input signal. For the 125 ns and  
6.425 μs filter settings, the jitter is up to 25 ns. On the 2.56 ms setting,  
the jitter is up to 10.025 μs.  
When a PFI input is routed directly to RTSI, or a RTSI input is routed  
directly to PFI, the s Series device does not use the filtered version of the  
input signal.  
Refer to the KnowledgeBase document, Digital Filtering with M Series  
and CompactDAQ, for more information about digital filters and counters.  
To access this KnowledgeBase, go to ni.com/infoand enter the info  
code rddfms.  
© National Instruments Corporation  
8-7  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 8  
Programmable Function Interfaces (PFI)  
I/O Protection  
Each DIO and PFI signal is protected against overvoltage, undervoltage,  
and overcurrent conditions as well as ESD events. However, you should  
avoid these fault conditions by following these guidelines:  
If you configure a PFI or DIO line as an output, do not connect it to any  
external signal source, ground, or power supply.  
If you configure a PFI or DIO line as an output, understand the current  
requirements of the load connected to these signals. Do not exceed the  
specified current output limits of the DAQ device. NI has several signal  
conditioning solutions for digital applications requiring high current  
drive.  
If you configure a PFI or DIO line as an input, do not drive the line with  
voltages outside of its normal operating range. The PFI or DIO lines  
have a smaller operating range than the AI signals.  
Treat the DAQ device as you would treat any static sensitive device.  
Always properly ground yourself and the equipment when handling the  
DAQ device or connecting to it.  
Programmable Power-Up States  
The programmable power-up state for the DAQ-STC2-based S Series  
devices are as follows:  
NI 6124 Devices— At system startup and reset, the hardware sets all  
PFI and DIO lines to high-impedance inputs by default. The DAQ  
device does not drive the signal high or low. Each line has a weak  
pull-down resistor connected to it, as described in the specifications  
document for your device.  
NI-DAQmx supports programmable power-up states for PFI and DIO  
lines. Software can program any value at power up to the P0, P1, or P2  
lines. The PFI and DIO lines can be set as:  
A high-impedance input with a weak pull-down resistor (default)  
An output driving a 0  
An output driving a 1  
Note When using your S Series device to control an SCXI chassis, DIO lines 0, 1, 2, and  
4 are used as communication lines and must be left to power-up in the default  
high-impedance state to avoid potential damage to these signals.  
NI 6124/6154 User Manual  
8-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 8  
Programmable Function Interfaces (PFI)  
NI 6154 Devices—By default, the digital output lines  
(P1.<0..3>/PFI <6..9>) are disabled (high impedance) at power up.  
Software can configure the board to power up with the entire port  
enabled or disabled; you cannot enable individual lines. If the port  
powers up enabled, you also can configure each line individually to  
power up as 1 or 0.  
Refer to the NI-DAQmx Help or the LabVIEW Help in version 8.0 or later  
for more information about setting power-up states in NI-DAQmx or MAX.  
© National Instruments Corporation  
8-9  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
9
Digital Routing and Clock  
Generation  
The digital routing circuitry has the following main functions:  
Manages the flow of data between the bus interface and the  
acquisition/generation sub-systems (analog input, analog output,  
digital I/O, and the counters). The digital routing circuitry uses FIFOs  
(if present) in each sub-system to ensure efficient data movement.  
Routes timing and control signals. The acquisition/generation  
sub-systems use these signals to manage acquisitions and generations.  
These signals can come from the following sources:  
Your S Series device  
Other devices in your system through RTSI  
User input through the PFI terminals  
User input through the PXI_STAR terminal  
Routes and generates the main clock signals for the S Series device.  
Clock Routing  
Figure 9-1 shows the clock routing circuitry of an S Series device.  
Onboard  
80 MHz  
Oscillator  
10 MHz RefClk  
(To RTSI <0..7>  
÷ 8  
Output Selectors)  
External  
Reference  
Clock  
80 MHz Timebase  
20 MHz Timebase  
RTSI <0..7>  
PXI_CLK10  
PLL  
÷ 4  
PXI_STAR  
100 kHz Timebase  
÷ 200  
Figure 9-1. S Series Clock Routing Circuitry  
© National Instruments Corporation  
9-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 9  
Digital Routing and Clock Generation  
80 MHz Timebase  
The 80 MHz Timebase can be used as the Source input to the 32-bit  
general-purpose counter/timers.  
The 80 MHz Timebase is generated from the following sources:  
Onboard oscillator  
External signal (by using the external reference clock)  
20 MHz Timebase  
100 kHz Timebase  
The 20 MHz Timebase normally generates many of the AI and AO timing  
signals. The 20 MHz Timebase also can be used as the Source input to the  
32-bit general-purpose counter/timers.  
The 20 MHz Timebase is generated by dividing down the 80 MHz  
Timebase.  
The 100 kHz Timebase can be used to generate many of the AI and AO  
timing signals. The 100 kHz Timebase also can be used as the Source input  
to the 32-bit general-purpose counter/timers.  
The 100 kHz Timebase is generated by dividing down the 20 MHz  
Timebase by 200.  
External Reference Clock  
The external reference clock can be used as a source for the internal  
timebases (80 MHz Timebase, 20 MHz Timebase, and 100 kHz Timebase)  
on an S Series device. By using the external reference clock, you can  
synchronize the internal timebases to an external clock.  
The following signals can be routed to drive the external reference clock:  
RTSI <0..7>  
PXI_CLK10  
PXI_STAR  
The external reference clock is an input to a Phase-Lock Loop (PLL). The  
PLL generates the internal timebases.  
NI 6124/6154 User Manual  
9-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
               
Chapter 9  
Digital Routing and Clock Generation  
10 MHz Reference Clock  
The 10 MHz reference clock can be used to synchronize other devices to  
your S Series device. The 10 MHz reference clock can be routed to the  
RTSI <0..7> terminals. Other devices connected to the RTSI bus can use  
this signal as a clock input.  
The 10 MHz reference clock is generated by dividing down the onboard  
oscillator.  
Synchronizing Multiple Devices  
With the RTSI bus and the routing capabilities of S Series devices, there are  
several ways to synchronize multiple devices depending on your  
application.  
To synchronize multiple devices to a common timebase, choose  
one device—the initiator—to generate the timebase. The initiator device  
routes its 10 MHz reference clock to one of the RTSI <0..7> signals.  
All devices (including the initiator device) receive the 10 MHz reference  
clock from RTSI. This signal becomes the external reference clock. A PLL  
on each device generates the internal timebases synchronous to the external  
reference clock.  
On PXI systems, you also can synchronize devices to PXI_CLK10. In this  
application the PXI chassis acts as the initiator. Each PXI module routes  
PXI_CLK10 to its external reference clock.  
Another option in PXI systems is to use PXI_STAR. The Star Trigger  
controller device acts as the initiator and drives PXI_STAR with a clock  
signal. Each target device routes PXI_STAR to its external reference clock.  
Once all of the devices are using or referencing a common timebase, you  
can synchronize operations across them by sending a common start trigger  
out across the RTSI bus and setting their sample clock rates to the same  
value.  
© National Instruments Corporation  
9-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 9  
Digital Routing and Clock Generation  
Real-Time System Integration (RTSI)  
Real-Time System Integration (RTSI) is a set of bused signals among  
devices that allows you to do the following:  
Use a common clock (or timebase) to drive the timing engine on  
multiple devices  
Share trigger signals between devices  
Many National Instruments DAQ, motion, vision, and CAN devices  
support RTSI.  
In a PCI system, the RTSI bus consists of the RTSI bus interface and a  
ribbon cable. The bus can route timing and trigger signals between several  
functions on as many as five DAQ, vision, motion, or CAN devices in the  
computer.  
In a PXI Express system, the RTSI bus consists of the RTSI bus interface  
and the PXI trigger signals on the PXI backplane. This bus can route timing  
and trigger signals between several functions on as many as seven DAQ  
devices in the system.  
RTSI Connector Pinout  
(NI 6154 Only) Figure 9-2 shows the RTSI connector pinout and Table 9-1  
describes the RTSI signals.  
Table 9-1. RTSI Signals  
Terminal  
RTSI Bus Signal  
1–18  
Not Connected. Do not connect  
signals to these terminals.  
19, 21, 23, 25, 27, 29, 31, 33  
D GND  
RTSI 0  
RTSI 1  
RTSI 2  
RTSI 3  
RTSI 4  
RTSI 5  
20  
22  
24  
26  
30  
NI 6124/6154 User Manual  
9-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 9  
Digital Routing and Clock Generation  
Table 9-1. RTSI Signals (Continued)  
Terminal RTSI Bus Signal  
32  
34  
RTSI 6  
RTSI 7  
Terminal 34  
Terminal 33  
Terminal 2  
Terminal 1  
Figure 9-2. S Series PCI Device RTSI Pinout  
Using RTSI as Outputs  
RTSI <0..7> are bidirectional terminals. As an output, you can drive any of  
the following signals to any RTSI terminal:  
AI Start Trigger (ai/StartTrigger)  
AI Reference Trigger (ai/ReferenceTrigger)  
AI Convert Clock* (ai/ConvertClock)  
AI Sample Clock (ai/SampleClock)  
AO Sample Clock* (ao/SampleClock)  
AO Start Trigger (ao/StartTrigger)  
AO Pause Trigger (ao/PauseTrigger)  
10 MHz Reference Clock  
© National Instruments Corporation  
9-5  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 9  
Digital Routing and Clock Generation  
Counter n Source, Gate, Z, Internal Output  
Change Detection Event  
Analog Comparison Event  
FREQ OUT  
PFI <0..5>  
Note Signals with a * are inverted before being driven on the RTSI terminals.  
Using RTSI Terminals as Timing Input Signals  
You can use RTSI terminals to route external timing signals to many  
different S Series functions. Each RTSI terminal can be routed to any of the  
following signals:  
AI Convert Clock (ai/ConvertClock)  
AI Sample Clock (ai/SampleClock)  
AI Start Trigger (ai/StartTrigger)  
AI Reference Trigger (ai/ReferenceTrigger)  
AI Sample Clock Timebase (ai/SampleClockTimebase)  
AO Start Trigger (ao/StartTrigger)  
AO Sample Clock (ao/SampleClock)  
AO Sample Clock Timebase (ao/SampleClockTimebase)  
AO Pause Trigger (ao/PauseTrigger)  
Counter input signals for either counter—Source, Gate, Aux,  
HW_Arm, A, B, or Z  
DI Sample Clock (di/SampleClock)  
DO Sample Clock (do/SampleClock)  
Most functions allow you to configure the polarity of PFI inputs and  
whether the input is edge or level sensitive.  
RTSI Filters  
You can enable a programmable debouncing filter on each PFI, RTSI, or  
PXI_STAR signal. When the filters are enabled, your device samples the  
input on each rising edge of a filter clock. S Series devices use an onboard  
oscillator to generate the filter clock with a 40 MHz frequency.  
Note NI-DAQmx only supports filters on counter inputs.  
NI 6124/6154 User Manual  
9-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 9  
Digital Routing and Clock Generation  
The following is an example of low to high transitions of the input signal.  
High to low transitions work similarly.  
Assume that an input terminal has been low for a long time. The input  
terminal then changes from low to high, but glitches several times. When  
the filter clock has sampled the signal high on N consecutive edges, the low  
to high transition is propagated to the rest of the circuit. The value of N  
depends on the filter setting; refer to Table 9-2.  
Table 9-2. Filters  
N (Filter Clocks  
Needed to  
Pass Signal)  
Pulse Width  
Guaranteed to  
Pass Filter  
Pulse Width  
Guaranteed to  
Not Pass Filter  
Filter Setting  
125 ns  
5
257  
125 ns  
6.425 μs  
2.56 ms  
100 ns  
6.400 μs  
2.54 ms  
6.425 μs  
2.56 ms  
~101,800  
Disabled  
The filter setting for each input can be configured independently. On power  
up, the filters are disabled. Figure 9-3 shows an example of a low to high  
transition on an input that has its filter set to 125 ns (N = 5).  
RTSI, PFI, or  
PXI_STAR Terminal  
Filtered input goes  
high when terminal  
is sampled high on  
five consecutive filter  
clocks.  
1
1
2
3
4
1
2
3
4
5
Filter Clock  
(40 MHz)  
Filtered Input  
Figure 9-3. Filter Example  
Enabling filters introduces jitter on the input signal. For the 125 ns and  
6.425 μs filter settings, the jitter is up to 25 ns. On the 2.56 ms setting,  
the jitter is up to 10.025 μs.  
When a PFI input is routed directly to RTSI, or a RTSI input is routed  
directly to PFI, the S Series device does not use the filtered version of  
the input signal.  
© National Instruments Corporation  
9-7  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 9  
Digital Routing and Clock Generation  
Refer to the KnowledgeBase document, Digital Filtering with M Series  
and CompactDAQ, for more information about digital filters and counters.  
To access this KnowledgeBase, go to ni.com/infoand enter the info  
code rddfms.  
PXI Clock and Trigger Signals  
(NI 6124 Only) PXI clock and trigger signals are only available on PXI and  
PXI Express devices.  
PXI_CLK10  
(NI 6124 Only) PXI_CLK10 is a common low-skew 10 MHz reference clock  
for synchronization of multiple modules in a PXI measurement or control  
system. The PXI backplane is responsible for generating PXI_CLK10  
independently to each peripheral slot in a PXI chassis.  
PXI Triggers  
(NI 6124 Only) A PXI chassis provides eight bused trigger lines to each  
module in a system. Triggers may be passed from one module to another,  
allowing precisely timed responses to asynchronous external events that  
are being monitored or controlled. Triggers can be used to synchronize the  
operation of several different PXI peripheral modules.  
On S Series devices, the eight PXI trigger signals are synonymous with  
RTSI <0..7>.  
Note that in a PXI chassis with more than eight slots, the PXI trigger lines  
may be divided into multiple independent buses. Refer to the  
documentation for your chassis for details.  
PXI_STAR Trigger  
(NI 6124 Only) In a PXI system, the Star Trigger bus implements a dedicated  
trigger line between the first peripheral slot (adjacent to the system slot)  
and the other peripheral slots. The Star Trigger can be used to synchronize  
multiple devices or to share a common trigger signal among devices.  
A Star Trigger controller can be installed in this first peripheral slot to  
provide trigger signals to other peripheral modules. Systems that do not  
require this functionality can install any standard peripheral module in this  
first peripheral slot.  
NI 6124/6154 User Manual  
9-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                     
Chapter 9  
Digital Routing and Clock Generation  
An S Series device receives the Star Trigger signal (PXI_STAR) from a  
Star Trigger controller. PXI_STAR can be used as an external source for  
many AI, AO, and counter signals.  
An S Series device is not a Star Trigger controller. An S Series device may  
be used in the first peripheral slot of a PXI system, but the system will not  
be able to use the Star Trigger feature.  
PXI_STAR Filters  
(NI 6124 Only) You can enable a programmable debouncing filter on each  
PFI, RTSI, or PXI_STAR signal. When the filters are enabled, your device  
samples the input on each rising edge of a filter clock. S Series devices use  
an onboard oscillator to generate the filter clock with a 40 MHz frequency.  
Note NI-DAQmx only supports filters on counter inputs.  
The following is an example of low to high transitions of the input signal.  
High to low transitions work similarly.  
Assume that an input terminal has been low for a long time. The input  
terminal then changes from low to high, but glitches several times. When  
the filter clock has sampled the signal high on N consecutive edges, the low  
to high transition is propagated to the rest of the circuit. The value of N  
depends on the filter setting; refer to Table 9-3.  
Table 9-3. Filters  
N (Filter Clocks  
Needed to  
Pass Signal)  
Pulse Width  
Guaranteed to  
Pass Filter  
Pulse Width  
Guaranteed to  
Not Pass Filter  
Filter Setting  
125 ns  
5
257  
125 ns  
6.425 μs  
2.56 ms  
100 ns  
6.400 μs  
2.54 ms  
6.425 μs  
2.56 ms  
~101,800  
Disabled  
© National Instruments Corporation  
9-9  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Chapter 9  
Digital Routing and Clock Generation  
The filter setting for each input can be configured independently. On power  
up, the filters are disabled. Figure 9-4 shows an example of a low to high  
transition on an input that has its filter set to 125 ns (N = 5).  
RTSI, PFI, or  
PXI_STAR Terminal  
Filtered input goes  
high when terminal  
is sampled high on  
five consecutive filter  
clocks.  
1
1
2
3
4
1
2
3
4
5
Filter Clock  
(40 MHz)  
Filtered Input  
Figure 9-4. Filter Example  
Enabling filters introduces jitter on the input signal. For the 125 ns and  
6.425 μs filter settings, the jitter is up to 25 ns. On the 2.56 ms setting,  
the jitter is up to 10.025 μs.  
When a PFI input is routed directly to RTSI, or a RTSI input is routed  
directly to PFI, the S Series device does not use the filtered version of the  
input signal.  
Refer to the KnowledgeBase document, Digital Filtering with M Series  
and CompactDAQ, for more information about digital filters and counters.  
To access this KnowledgeBase, go to ni.com/infoand enter the info  
code rddfms.  
Routing Signals in Software  
Table 9-4 lists the basic functions you can use to route signals.  
Table 9-4. Signal Routing in Software  
Function  
Language  
LabVIEW and NI-DAQmx  
DAQmx Export Signal.viand  
DAQmx Connect Terminals.vi  
C and NI-DAQmx  
Export_Signaland DAQmx_Connect_Terminals  
Note For more information about routing signals in software, refer to the NI-DAQmx Help  
or the LabVIEW Help in version 8.0 or later.  
NI 6124/6154 User Manual  
9-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
10  
Bus Interface  
Each S Series device is designed on a complete hardware architecture that  
is deployed on the following platforms:  
PCI  
PXI Express  
Using NI-DAQmx driver software, you have the flexibility to change  
hardware platforms and operating systems with little or no change to  
software code.  
MITE and DAQ-PnP  
All S Series devices are jumperless for complete plug-and-play operation.  
The operating system automatically assigns the base address, interrupt  
levels, and other resources.  
NI S Series PCI/PXIe devices incorporate PCI-MITE technology to  
implement a high-performance PCI interface.  
PXI Considerations  
PXI Clock and Trigger Signals  
(NI 6124 Only) Refer to the PXI_CLK10, PXI Triggers, PXI_STAR Trigger,  
and PXI_STAR Filters sections of Chapter 9, Digital Routing and Clock  
Generation, for more information about PXI clock and trigger signals.  
PXI Express  
(NI 6124 Only) NI PXI Express S Series devices can be installed in any PXI  
Express slot or PXI hybrid slots in PXI Express chassis. PXI specifications  
are developed by the PXI System Alliance (www.pxisa.org).  
© National Instruments Corporation  
10-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                       
Chapter 10  
Bus Interface  
Data Transfer Methods  
There are three primary ways to transfer data across the PCI bus are as  
follows:  
Direct Memory Access (DMA)—DMA is a method to transfer data  
between the device and computer memory without the involvement of  
the CPU. This method makes DMA the fastest available data transfer  
method. National Instruments uses DMA hardware and software  
technology to achieve high throughput rates and to increase system  
utilization. DMA is the default method of data transfer for DAQ  
devices that support it.  
Interrupt Request (IRQ)—IRQ transfers rely on the CPU to service  
data transfer requests. The device notifies the CPU when it is ready to  
transfer data. The data transfer speed is tightly coupled to the rate at  
which the CPU can service the interrupt requests. If you are using  
interrupts to acquire data at a rate faster than the rate the CPU can  
service the interrupts, your systems may start to freeze.  
Programmed I/O—Programmed I/O is a data transfer mechanism  
where the user’s program is responsible for transferring data. Each  
read or write call in the program initiates the transfer of data.  
Programmed I/O is typically used in software-timed (on demand)  
operations.  
Changing Data Transfer Methods between DMA and IRQ  
There are a limited number of DMA channels per device (refer to the  
specifications document for your device). Each operation (specifically, AI,  
AO, and so on) that requires a DMA channel uses that method until all of  
the DMA channels are used. After all of the DMA channels are used, you  
will get an error if you try to run another operation requesting a DMA  
channel. If appropriate, you can change one of the operations to use  
interrupts. For NI-DAQmx, use the Data Transfer Mechanism property  
node.  
NI 6124/6154 User Manual  
10-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
11  
Triggering  
A trigger is a signal that causes a device to perform an action, such as  
starting an acquisition. You can program your DAQ device to generate  
triggers on any of the following:  
A software command  
A condition on an external digital signal  
A condition on an external analog signal  
You can also program your DAQ device to perform an action in response to  
a trigger. The action can affect the following:  
Analog input acquisitions  
Analog output generation  
Counter behavior  
For more information about analog input triggering, refer to Chapter 4,  
Analog Input. For more information about analog output triggering, refer to  
Chapter 5, Analog Output. For more information about counter triggering,  
refer to Chapter 7, Counters.  
Note (NI 6154 Only) NI 6154 devices do not support analog triggering. For information  
about the triggering capabilities of your device, refer to the specifications document for  
your device.  
Triggering with a Digital Source  
Your DAQ device can generate a trigger on a digital signal. You must  
specify a source and an edge. The digital source can be any of the input  
PFIs or RTSI <0..6> signals.  
The edge can be either the rising edge or falling edge of the digital signal.  
A rising edge is a transition from a low logic level to a high logic level.  
A falling edge is a high to low transition.  
© National Instruments Corporation  
11-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Chapter 11  
Triggering  
Figure 11-1 shows a falling-edge trigger.  
5 V  
Digital Trigger  
0 V  
Falling edge initiates acquisition  
Figure 11-1. Falling-Edge Trigger  
You can also program your DAQ device to perform an action in response to  
a trigger from a digital source. The action can affect the following:  
analog input acquisitions  
analog output generation  
counter behavior  
Triggering with an Analog Source  
(NI 6124 Only) Some S Series devices can generate a trigger on an analog  
signal. Figure 11-2 shows the analog trigger circuitry.  
ADC  
+
AI 0  
AI 1  
PGIA  
ADC  
+
PGIA  
AI Circuitry  
Analog  
Trigger  
Detection  
Analog Comparison Event  
(Analog Trigger Circuitry Output)  
AO Circuitry  
Mux  
Counter Circuitry  
ADC  
ADC  
+
AI 2  
AI 3  
PGIA  
+
PGIA  
Figure 11-2. Analog Trigger Circuitry  
You must specify a source and an analog trigger type. The source can be  
any analog input channel.  
NI 6124/6154 User Manual  
11-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Chapter 11  
Triggering  
Analog Input Channel  
instrumentation amplifier. The instrumentation amplifier amplifies the  
signal as determined by the input mode and the input polarity and range.  
The output of the instrumentation amplifier then drives the analog trigger  
detection circuit. By using the instrumentation amplifier, you can trigger on  
very small voltage changes in the input signal. For more information, refer  
to the Analog Trigger Accuracy section.  
Analog Trigger Actions  
(NI 6124 Only) The output of the Analog Trigger Detection circuit is the  
Analog Comparison Event signal. You can program your DAQ device to  
perform an action in response to the Analog Comparison Event signal. The  
action can affect the following:  
Analog input acquisitions  
Analog output generation  
Counter behavior  
Note Refer to Timing and Triggering in the NI-DAQmx Help or the LabVIEW Help in  
version 8.0 or later for more information.  
Note (NI 6154 Only) The NI 6154 does not support hardware analog triggering.  
Analog Trigger Types  
(NI 6124 Only) Configure the analog trigger circuitry to different triggering  
modes:  
Analog Edge Triggering—Configure the analog trigger circuitry to  
detect when the analog signal is below or above a level you specify.  
© National Instruments Corporation  
11-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Chapter 11  
Triggering  
In below-level analog triggering mode, shown in Figure 11-3, the  
trigger is generated when the signal value is less than Level.  
Level  
Analog Comparison Event  
Figure 11-3. Below-Level Analog Triggering Mode  
In above-level analog triggering mode, shown in Figure 11-4, the  
trigger is generated when the signal value is greater than Level.  
Level  
Analog Comparison Event  
Figure 11-4. Above-Level Analog Triggering Mode  
Analog Edge Triggering with Hysteresis—Hysteresis adds a  
programmable voltage region above or below the trigger level that an  
input signal must pass through before the DAQ device recognizes a  
trigger condition, and is often used to reduce false triggering due to  
noise or jitter in the signal.  
Analog Edge Trigger with Hysteresis (Rising Slope)—When  
using hysteresis with a rising slope, you specify a trigger level and  
amount of hysteresis. The high threshold is the trigger level; the  
low threshold is the trigger level minus the hysteresis.  
NI 6124/6154 User Manual  
11-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Chapter 11  
Triggering  
For the trigger to assert, the signal must first be below the low  
threshold, then go above the high threshold. The trigger stays  
asserted until the signal returns below the low threshold. The  
output of the trigger detection circuitry is the internal Analog  
Comparison Event signal, as shown in Figure 11-5.  
Then signal must  
go above high threshold before  
Analog Comparison Event asserts  
High threshold  
(Level)  
Hysteresis  
Low threshold  
(Level – Hysteresis)  
First signal must go  
below low threshold  
Analog Comparison Event  
Figure 11-5. Analog Edge Triggering with Hysteresis Rising Slope Example  
Analog Edge Trigger with Hysteresis (Falling Slope)—When  
using hysteresis with a falling slope, you specify a trigger level  
and amount of hysteresis. The low threshold is the trigger level;  
the high threshold is the trigger level plus the hysteresis.  
For the trigger to assert, the signal must first be above the high  
threshold, then go below the low threshold. The trigger stays  
asserted until the signal returns above the high threshold. The  
output of the trigger detection circuitry is the internal Analog  
Comparison Event signal, as shown in Figure 11-6.  
First signal must go  
above high threshold  
High threshold  
(Level + Hysteresis)  
Hysteresis  
Low threshold  
(Level)  
Then signal must go below low threshold  
before Analog Comparison Event asserts  
Analog Comparison Event  
Figure 11-6. Analog Edge Triggering with Hysteresis Falling Slope Example  
© National Instruments Corporation  
11-5  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Chapter 11  
Triggering  
Analog Window Triggering—An analog window trigger occurs  
when an analog signal either passes into (enters) or passes out of  
(leaves) a window defined by two voltage levels. Specify the levels by  
setting the window Top value and the window Bottom value.  
Figure 11-7 demonstrates a trigger that asserts when the signal enters  
the window.  
Top  
Bottom  
Analog Comparison Event  
Figure 11-7. Analog Window Triggering Mode (Entering Window)  
Analog Trigger Accuracy  
(NI 6124 Only) The analog trigger circuitry compares the voltage of the  
trigger source to the output of programmable trigger DACs. When you  
configure the level (or the high and low limits in window trigger mode), the  
device adjusts the output of the trigger DACs. Refer to the specifications  
document for your device to find the accuracy and resolution of the analog  
trigger DACs.  
To improve accuracy you can software-calibrate the analog trigger  
circuitry. No hardware calibration is provided for the analog trigger  
circuitry. In addition, the propagation delay from when a valid trigger  
condition is met to when the analog trigger circuitry emits the Analog  
Comparison Event may have an impact on your measurements if the trigger  
signal has a high slew rate. If you find these conditions have a noticeable  
impact on your measurements, you can perform software calibration on the  
analog trigger circuitry by configuring your task as normal and applying a  
known signal for your analog trigger. Comparing the observed results  
against the expected results, you can calculate the necessary offsets to apply  
in software to fine-tune the desired triggering behavior.  
NI 6124/6154 User Manual  
11-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
A
Device-Specific Information  
This appendix includes device-specific information about the following  
S Series devices:  
NI 6124  
NI 6154  
NI 6124  
The NI 6124 is a Plug-and-Play, multifunction analog, digital, and timing  
I/O device for PXI Express bus computers.  
The NI 6124 features:  
Four simultaneously sampling analog inputs at 4 MS/s with one 16-bit  
A/D converter (ADC) per channel  
Two 16-bit D/A converters (DACs) with voltage outputs  
Eight lines of TTL-compatible correlated DIO  
16 lines of TTL-compatible static DIO  
Two general-purpose 32-bit counter/timers  
Increased common-mode noise rejection through differential signal  
connection  
Because the NI 6124 has no DIP switches, jumpers, or potentiometers, it  
can be easily calibrated and configured in software.  
NI 6124 Analog Output  
The NI 6124 supplies two channels of AO voltage at the I/O connector.  
The range is fixed at bipolar 10 V.  
The AO channels on the NI 6124 contain 16-bit DACs that are capable of  
4 MS/s for one channel or 2.5 MS/s for each of two channels. Refer to the  
NI 6124 Specifications for more detailed information about the  
AO capabilities of the NI 6124.  
© National Instruments Corporation  
A-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Appendix A  
Device-Specific Information  
Note The AO channels do not have analog or digital filtering hardware and do produce  
images in the frequency domain related to the update rate.  
NI 6124 I/O Connector Pinout  
Figure A-1 shows the pin assignments for the 68-pin connector on the  
NI 6124.  
68 34  
67 33  
66 32  
65 31  
64 30  
63 29  
62 28  
61 27  
60 26  
59 25  
58 24  
57 23  
56 22  
55 21  
54 20  
53 19  
52 18  
51 17  
50 16  
49 15  
48 14  
47 13  
46 12  
45 11  
44 10  
AI 0 +  
AI 0 –  
AI 1 +  
AI 1 GND  
AI 2 –  
AI 3 +  
AI 3 GND  
NC  
AI 0 GND  
AI 1 –  
AI 2 +  
AI 2 GND  
AI 3 –  
NC  
NC  
NC  
NC  
NC  
NC  
NC  
NC  
NC  
NC  
NC  
NC  
AO 0  
AO GND  
AO GND  
D GND  
P0.0  
AO 1  
NC  
P0.4  
D GND  
P0.1  
P0.5  
D GND  
P0.2  
P0.6  
D GND  
+5 V  
P0.7  
D GND  
D GND  
PFI 0/P1.0  
PFI 1/P1.1  
D GND  
+5 V  
P0.3  
PFI 11/P2.3  
PFI 10/P2.2  
D GND  
PFI 2/P1.2  
PFI 3/P1.3  
PFI 4/P1.4  
PFI 13/P2.5  
PFI 15/P2.7  
PFI 7/P1.7  
PFI 8/P2.0  
D GND  
D GND  
43  
42  
41  
40  
39  
38  
37  
36  
35  
9
8
7
6
5
4
3
2
1
D GND  
PFI 5/P1.5  
PFI 6/P1.6  
D GND  
PFI 9/P2.1  
PFI 12/P2.4  
PFI 14/P2.6  
NC = No Connect  
Figure A-1. NI 6124 Pinout  
NI 6124/6154 User Manual  
A-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Appendix A  
Device-Specific Information  
Table A-1. Default NI-DAQmx Counter/Timer Pins  
Counter/Timer Signal  
CTR 0 SRC  
Default Pin Number (Name)  
37 (PFI 8)  
CTR 0 GATE  
CTR 0 AUX  
CTR 0 OUT  
CTR 0 A  
3 (PFI 9)  
45 (PFI 10)  
2 (PFI 12)  
37 (PFI 8)  
3 (PFI 9)  
CTR 0 Z  
CTR 0 B  
45 (PFI 10)  
42 (PFI 3)  
41 (PFI 4)  
46 (PFI 11)  
40 (PFI 13)  
42 (PFI 3)  
41 (PFI 4)  
46 (PFI 11)  
1 (PFI 14)  
CTR 1 SRC  
CTR 1 GATE  
CTR 1 AUX  
CTR 1 OUT  
CTR 1 A  
CTR 1 Z  
CTR 1 B  
FREQ OUT  
Note For more information about default NI-DAQmx counter inputs, refer to Connecting  
Counter Signals in the NI-DAQmx Help or the LabVIEW Help in version 8.0 or later.  
For a detailed description of each signal, refer to the NI 6124 I/O Connector  
Signal Descriptions section of Chapter 3, I/O Connector.  
© National Instruments Corporation  
A-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
Device-Specific Information  
NI 6124 Block Diagram  
Figure A-2 shows the NI 6124 block diagram.  
AI Data ADC(0)  
16-Bit  
PGIA  
PGIA  
PGIA  
SAR  
Channel  
Cal Relay  
Control(0)  
AI Convert(0)  
Control(0)  
AI Data ADC(1)  
16-Bit  
SAR  
Channel  
Control(1)  
Board  
Power  
Cal Relay  
Control(1)  
AI Convert(1)  
Spartan-3A  
AI Data ADC(2)  
16-Bit  
SAR  
FPGA  
Channel  
Control(2)  
Cal Relay  
Control(2)  
AI Convert(2)  
3.3V  
+12V  
AI Data ADC(3)  
16-Bit  
SAR  
PGIA  
Channel  
Control(3)  
Cal Relay  
Control(3)  
AI Convert(3)  
Temp  
Sensor  
Calibration Control  
Config  
CPLD  
Convert Clock  
Buffer  
Calibration  
PWM  
Flash  
Calibration  
Multiplexer  
Mite I/O Port  
Voltage  
Reference  
AO Control DAC (0)  
AI Convert  
X1 Link  
PCIe  
to PCI  
Bridge  
AI  
AO Data DAC (0)  
16-Bit  
DAC 0  
DAQ-STC2  
AO Control DAC (1)  
AO  
Address/  
ASIC  
PCI  
Port  
Data  
AO Data DAC (1)  
16-Bit  
DAC 1  
Digital I/O (8)  
Digital I/O  
PFI  
DAQ-6202  
ASIC  
(Lightning  
Bug)  
DIO  
DIO  
Protection  
PFI (16)  
Configure  
Digital I/O  
PFI  
EEPROM  
Figure A-2. NI 6124 Block Diagram  
NI 6124 Cables and Accessories  
This section describes some of the cable and accessory options for the  
NI 6124. For more specific information about these products, refer to  
ni.com.  
NI 6124/6154 User Manual  
A-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Appendix A  
Device-Specific Information  
Using BNCs  
You can connect BNC cables to your DAQ device using BNC accessories  
such as the BNC-2110, BNC-2120, and BNC-2090A. To connect your  
DAQ device to a BNC accessory, use one of the following cables:  
SH68-68-EPM—shielded cable  
SH68-68R1-EP—shielded cable with one right angle connector  
SH6868—shielded 68-conductor cable  
RC68-68—unshielded cable  
Using Screw Terminals  
You can connect signals to your DAQ device using a screw terminal  
accessory such as:  
CB-68LP, CB-68LPR—low-cost screw terminal block  
SCB-68—shielded screw terminal block with breadboard areas  
TBX-68—DIN rail mountable screw terminal block  
To connect your DAQ device to a screw terminal accessory, use one of the  
following cables:  
SH68-68-EPM—shielded cable  
SH68-68R1-EP—shielded cable with one right angle connector  
RC68-68—unshielded cable  
Using SSR or ER Digital Signal Conditioning  
SSR and ER series provide per channel digital signal conditioning.  
Custom Cabling/Connectors Options  
The CA-1000 is a versatile connector/enclosure system. It allows the user  
to define I/O connectors on a per-channel basis. Internally, the system  
allows for flexible custom wiring configuration.  
If you want to develop your own cable, follow these guidelines for best  
results:  
Use shielded twisted-pair wires for each differential AI pair. Connect  
the shield for each signal pair to the ground reference at the source.  
Route the analog lines separately from the digital lines.  
When using a cable shield, use separate shields for the analog and  
digital halves of the cable. Failure to do so results in noise coupling  
into the analog signals from transient digital signals.  
© National Instruments Corporation  
A-5  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
Device-Specific Information  
Mating connectors and a backshell kit for making custom 68-pin cables are  
available from NI.  
NI recommends that you use one of the following connectors with the I/O  
connector on your device.  
Honda 68-position, solder cup, female connector  
Honda backshell  
AMP VHDCI connector  
For more information about the connectors used for DAQ devices, refer to  
the KnowledgeBase document, Specifications and Manufacturers for  
Board Mating Connectors, by going to ni.com/infoand entering the info  
code rdspmb.  
NI 6124 Specifications  
Refer to the NI 6124 Specifications for more detailed information about the  
device.  
NI 6124/6154 User Manual  
A-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix A  
Device-Specific Information  
NI 6154  
The NI 6154 is an isolated Plug-and-Play multifunction analog, digital, and  
timing I/O device for PCI bus computers.  
The NI 6154 features:  
Four differential 16-bit analog inputs  
Four 16-bit analog output channels  
Ten lines of DIO (6 DI and 4 DO)  
Isolation for each AI and AO channel from the chassis and from one  
another  
Bank isolation for DIO from the chassis  
Because the NI 6154 has no DIP switches, jumpers, or potentiometers, it  
can be easily calibrated and configured in software.  
NI 6154 Analog Output  
The NI 6154 supplies four channels of AO voltage at the I/O connector. The  
range is fixed at bipolar 10 V.  
The AO channels on the NI 6154 contain 16-bit DACs that are capable of  
250 kS/s for each channel. The AO channels are isolated from each other,  
from the AI channels, and from the chassis. Refer to the NI 6154  
Specifications for more detailed information about the AO capabilities of  
the NI 6154.  
© National Instruments Corporation  
A-7  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Appendix A  
Device-Specific Information  
NI 6154 I/O Connector Pinout  
Figure A-3 shows the pin assignments for the 37-pin I/O connector on the  
NI 6154.  
1
2
AI 0–  
AI 0+  
20  
21  
22  
23  
24  
25  
26  
27  
28  
29  
30  
31  
32  
33  
34  
35  
36  
37  
NC  
AI 1–  
3
AI 1+  
NC  
4
AI 2–  
AI 2+  
5
NC  
AI 3–  
6
AI 3+  
NC  
7
AO 0–  
AO 0+  
8
NC  
AO 1–  
9
AO 1+  
NC  
10  
11  
12  
13  
14  
15  
16  
17  
18  
19  
AO 2–  
AO 2+  
NC  
AO 3–  
AO 3+  
NC  
PFI 0/P0.0 (Input)  
D GND  
PFI 1/P0.1 (Input)  
PFI 2/P0.2 (Input)  
PFI 4/P0.4 (Input)  
PFI 5/P0.5 (Input)  
PFI 7/P1.1 (Output)  
PFI 8/P1.2 (Output)  
PFI 3/P0.3 (Input)  
D GND  
PFI 6/P1.0 (Output)  
D GND  
PFI 9/P1.3 (Output)  
NC = No Connect  
Figure A-3. NI 6154 Pinout  
Table A-2. NI 6154 Device Default NI-DAQmx Counter/Timer Pins  
Counter/Timer Signal  
CTR 0 SRC  
CTR 0 GATE  
CTR 0 AUX  
CTR 0 OUT  
CTR 0 A  
Default Pin Number (Name)  
13 (PFI 0)  
Port  
P0.0  
P0.1  
P0.2  
P1.0  
P0.0  
P0.1  
32 (PFI 1)  
33 (PFI 2)  
17 (PFI 6)  
13 (PFI 0)  
CTR 0 Z  
32 (PFI 1)  
NI 6124/6154 User Manual  
A-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Appendix A  
Device-Specific Information  
Table A-2. NI 6154 Device Default NI-DAQmx Counter/Timer Pins (Continued)  
Counter/Timer Signal  
CTR 0 B  
Default Pin Number (Name)  
33 (PFI 2)  
Port  
P0.2  
P0.3  
P0.4  
P0.5  
P1.1  
P0.3  
P0.4  
CTR 1 SRC  
CTR 1 GATE  
CTR 1 AUX  
CTR 1 OUT  
CTR 1 A  
15 (PFI 3)  
34 (PFI 4)  
35 (PFI 5)  
36 (PFI 7)  
15 (PFI 3)  
CTR 1 Z  
34 (PFI 4)  
35 (PFI 5)  
Note For more information about default NI-DAQmx counter inputs, refer to Connecting  
Counter Signals in the NI-DAQmx Help or the LabVIEW Help in version 8.0 or later.  
For a detailed description of each signal, refer to the NI 6154 I/O Connector  
Signal Descriptions section of Chapter 3, I/O Connector.  
© National Instruments Corporation  
A-9  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Appendix A  
Device-Specific Information  
NI 6154 Block Diagram  
Figure A-4 shows the NI 6154 block diagram.  
CAL  
Circuit  
AI 0  
16-Bit  
ADC  
CAL  
MUX  
PGIA  
FPGA  
AO 0  
AO 0  
16-Bit  
DAC  
PGIA  
AI 1  
16-Bit  
ADC  
CAL  
MUX  
PGIA  
AO 1  
AO 1  
16-Bit  
DAC  
PGIA  
Analog  
Output  
Timing/  
Control  
Analog  
Input  
Timing/  
Control  
Bus  
Interface  
AI 2  
16-Bit  
ADC  
CAL  
MUX  
PGIA  
DMA/  
Interrupt  
Interface  
Trigger DAQ-  
Interface  
STC2  
AO 2  
AO 2  
16-Bit  
DAC  
PGIA  
Counter/  
Digital I/O  
Timing I/O  
RTSI Bus  
Interface  
AI 3  
16-Bit  
ADC  
CAL  
MUX  
PGIA  
AO 3  
AO 3  
16-Bit  
DAC  
PGIA  
DIO  
Voltage  
Regulation  
Figure A-4. NI 6154 Block Diagram  
NI 6154 Cables and Accessories  
This section describes some of the cable and accessory options for the  
NI 6154. For more specific information about these products, refer to  
ni.com.  
Using Screw Terminals  
You can connect signals to your DAQ device using a screw terminal  
accessory such as:  
CB-37F-LP—unshielded, I/O connector block with 37-pin D-SUB  
NI 6124/6154 User Manual  
A-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Appendix A  
Device-Specific Information  
CB-37FH—37-pin screw terminal block, horizontal, DIN rail mount  
CB-37FV—37-pin screw terminal block, vertical, DIN rail mount  
TB-37F-37CP—37-pin crimp & poke terminals, shell with strain  
relief  
TB-37F-37SC—37-pin solder cup terminals, shell with strain relief  
CB-37F-HVD—37-pin screw terminal block, 150 V CAT II, DIN rail  
mount  
To connect your DAQ device to a screw terminal accessory, use one of the  
following cables:  
SH37F-37M-2—37-pin female-to-male shielded I/O cable, 150 V,  
2 m (non-EMI shielding)  
SH37F-37M-1—37-pin female-to-male shielded I/O cable, 150 V,  
1 m (non-EMI shielding)  
SH37F-P-4—37-pin female-to-pigtails shielded I/O cable, 4 m  
R37F-37M-1—37-pin female-to-male ribbon I/O cable, 1 m1  
DB37M-DB37F-EP—37-pin male-to-female enhanced performance  
shielded I/O cable, 1 m (EMI shielding)  
For more information about optional equipment available from National  
Instruments Web site at ni.com.  
NI 6154 Isolation and Digital Isolators  
The NI 6154 is an isolated data acquisition device. As shown in Figure 2-2,  
General NI 6124 Block Diagram, the analog input, analog output, and  
PFI/static DIO circuitry are referenced to separate isolated grounds for each  
circuit. The bus interface circuitry, RTSI, digital routing, and clock  
generation are all referenced to a non-isolated ground. Table A-3 shows the  
ground symbols referenced in Figure 2-2.  
Table A-3. Ground Symbols  
Ground  
Isolated Ground  
Symbol  
Non-Isolated Ground  
1
Not recommended for EMC-sensitive applications.  
© National Instruments Corporation  
A-11  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Appendix A  
Device-Specific Information  
The non-isolated ground is connected to the chassis ground of the PC or  
chassis where the device is installed.  
The isolated ground is not connected to the chassis ground of the PC or  
chassis. The isolated ground can be at a higher or lower voltage relative to  
the non-isolated ground. All analog measurements are made relative to the  
connect this ground to the ground of the system being measured or  
controlled. For more information, refer to the following:  
Input  
Figure 4-2, Isolated S Series Analog Input Block Diagram  
Output  
Figure 5-2, Isolated S Series Device Analog Output Block Diagram  
The Connecting Digital I/O Signals on Isolated Devices section of  
Chapter 6, Digital I/O  
Figure 6-5, Isolated S Series Devices Digital I/O Block Diagram  
Chapter 8, Programmable Function Interfaces (PFI)  
Digital Isolation  
The NI 6154 uses digital isolators. Unlike analog isolators, digital isolators  
do not introduce any analog error in the measurements taken by the device.  
The A/D converter, used for analog input, is on the isolated side of the  
device. The analog inputs are digitized before they are sent across the  
isolation barrier. Similarly, the D/A converters, used for analog output, are  
on the isolated side of the device.  
Benefits of an Isolated DAQ Device  
With isolation, engineers can safely measure a small voltage in the presence  
of a large common-mode signal. Some advantages of isolation are as  
follows:  
Improved rejection—Isolation increases the ability of the  
measurement system to reject common-mode voltages.  
Common-mode voltage is the signal that is present or “common” to  
both the positive and negative input of a measurement device, but is not  
part of the signal to be measured.  
NI 6124/6154 User Manual  
A-12  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix A  
Device-Specific Information  
Improved accuracy—Isolation improves measurement accuracy by  
physically preventing ground loops. Ground loops, a common source  
of error and noise, are the result of a measurement system having  
multiple grounds at different potentials.  
Improved safety—Isolation creates an insulation barrier so you can  
make floating measurements while protecting against large transient  
voltage spikes.  
NI 6154 Specifications  
Refer to the NI 6154 Specifications for more detailed information about the  
device.  
© National Instruments Corporation  
A-13  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
B
Technical Support and  
Professional Services  
Visit the following sections of the award-winning National Instruments  
Web site at ni.comfor technical support and professional services:  
Support—Technical support at ni.com/supportincludes the  
following resources:  
Self-Help Technical Resources—For answers and solutions,  
visit ni.com/supportfor software drivers and updates,  
a searchable KnowledgeBase, product manuals, step-by-step  
troubleshooting wizards, thousands of example programs,  
tutorials, application notes, instrument drivers, and so on.  
Registered users also receive access to the NI Discussion Forums  
at ni.com/forums. NI Applications Engineers make sure every  
question submitted online receives an answer.  
Standard Service Program Membership—This program  
entitles members to direct access to NI Applications Engineers  
via phone and email for one-to-one technical support as well as  
exclusive access to on demand training modules via the Services  
Resource Center. NI offers complementary membership for a full  
year after purchase, after which you may renew to continue your  
benefits.  
For information about other technical support options in your  
area, visit ni.com/services, or contact your local office at  
ni.com/contact.  
Training and Certification—Visit ni.com/trainingfor  
self-paced training, eLearning virtual classrooms, interactive CDs,  
and Certification program information. You also can register for  
instructor-led, hands-on courses at locations around the world.  
System Integration—If you have time constraints, limited in-house  
technical resources, or other project challenges, National Instruments  
Alliance Partner members can help. To learn more, call your local  
NI office or visit ni.com/alliance.  
© National Instruments Corporation  
B-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                   
Appendix B  
Technical Support and Professional Services  
Declaration of Conformity (DoC)—A DoC is our claim of  
compliance with the Council of the European Communities using  
the manufacturer’s declaration of conformity. This system affords  
the user protection for electromagnetic compatibility (EMC) and  
product safety. You can obtain the DoC for your product by visiting  
ni.com/certification.  
Calibration Certificate—If your product supports calibration,  
you can obtain the calibration certificate for your product at  
ni.com/calibration.  
If you searched ni.comand could not find the answers you need, contact  
your local office or NI corporate headquarters. Phone numbers for our  
worldwide offices are listed at the front of this manual. You also can visit  
the Worldwide Offices section of ni.com/niglobalto access the branch  
office Web sites, which provide up-to-date contact information, support  
phone numbers, email addresses, and current events.  
NI 6124/6154 User Manual  
B-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Glossary  
Symbol  
Prefix  
pico  
Value  
10–12  
10–9  
10–6  
10–3  
103  
p
n
nano  
micro  
milli  
kilo  
μ
m
k
M
mega  
106  
Symbols  
°
Degree.  
>
<
Greater than.  
Less than.  
Negative of, or minus.  
Ohms.  
Ω
/
Per.  
%
Percent.  
Plus or minus.  
Positive of, or plus.  
+
A
A
Amperes—the unit of electric current.  
Analog-to-digital.  
A/D  
AC  
Alternating current.  
© National Instruments Corporation  
G-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Glossary  
ADC  
ADE  
Analog-to-digital converter—an electronic device, often an integrated  
circuit, that converts an analog voltage to a digital number.  
Application Development Environment—a software environment  
incorporating the development, debug, and analysis tools for software  
development. LabVIEW, Measurement Studio, and Visual Studio are  
examples.  
AI  
1. Analog input.  
2. Analog input channel signal.  
aliasing  
The consequence of sampling that causes signals with frequencies higher  
than half the sampling frequency to appear as lower frequency components  
in a frequency spectrum.  
AO  
Analog output.  
ASIC  
Application-Specific Integrated Circuit—a proprietary semiconductor  
component designed and manufactured to perform a set of specific  
functions.  
B
bipolar  
A signal range that includes both positive and negative values (for example,  
–5 to +5 V).  
building ground  
See earth ground.  
C
C
Celsius  
CalDAC  
Calibration DAC.  
NI 6124/6154 User Manual  
G-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Glossary  
channel  
1. Physical—a terminal or pin at which you can measure or generate an  
analog or digital signal. A single physical channel can include more than  
one terminal, as in the case of a differential analog input channel or a digital  
port of eight lines. The name used for a counter physical channel is an  
exception because that physical channel name is not the name of the  
terminal where the counter measures or generates the digital signal.  
2. Virtual—a collection of property settings that can include a name, a  
physical channel, input terminal connections, the type of measurement or  
generation, and scaling information. You can define NI-DAQmx virtual  
channels outside a task (global) or inside a task (local). Configuring virtual  
channels is integral to every measurement you take in NI-DAQmx. In  
NI-DAQmx, you can configure virtual channels either in MAX or in a  
program, and you can configure channels as part of a task or separately.  
chassis ground  
Any connection back to the protective conductor earth ground. See also  
earth ground.  
cm  
Centimeter.  
CMOS  
CMRR  
Complementary metal-oxide semiconductor.  
Common-mode rejection ratio—a measure of the capability of an  
instrument to reject a signal that is common to both input leads.  
CompactPCI  
A Eurocard configuration of the PCI bus for industrial applications.  
correlated DIO  
A feature that allows you to clock digital I/O on the same clock as  
analog I/O.  
counter/timer  
coupling  
A circuit that counts external pulses or clock pulses (timing).  
The manner in which a signal is connected from one circuit to another.  
When applied to instrument products or DAQ cards, it refers to the input  
signal coupling technique.  
CTR  
Counter signal.  
© National Instruments Corporation  
G-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Glossary  
D
D/A  
Digital-to-analog.  
DAC  
Digital-to-analog converter—an electronic device, often an integrated  
circuit, that converts a digital number into a corresponding analog voltage  
or current.  
DAQ  
See data acquisition (DAQ).  
DAQ device  
A device that acquires or generates data and can contain multiple channels  
and conversion devices. DAQ devices include plug-in devices, PCMCIA  
cards, and DAQPad devices, which connect to a computer USB or 1394  
(FireWire)® port. SCXI modules are considered DAQ devices.  
DAQ-STC  
Data acquisition system timing controller—an application-specific  
integrated circuit (ASIC) for the system timing requirements of a general  
A/D and D/A system.  
data acquisition  
(DAQ)  
1. Acquiring and measuring analog or digital electrical signals from  
sensors, transducers, and test probes or fixtures.  
2. Generating analog or digital electrical signals.  
dB  
Decibel—the unit for expressing a logarithmic measure of the ratio of two  
signal levels: dB = 20log10 V1/V2 for signals in volts.  
dBc  
DC  
Decibel carrier—level difference referenced to a carrier level, c.  
Direct current—although the term speaks of current, many different types  
of DC measurements are made, including DC Voltage, DC current, and DC  
power.  
device  
1. An instrument or controller you can access as a single entity that controls  
or monitors real-world I/O points. A device often is connected to a host  
computer through some type of communication network.  
2. See DAQ device.  
DI  
Digital input.  
DIFF  
Differential mode—an analog input mode consisting of two terminals, both  
of which are isolated from computer ground, whose difference is measured.  
DIO  
Digital input/output.  
NI 6124/6154 User Manual  
G-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
DIP  
Dual inline package.  
DMA  
Direct memory access—a method by which data can be transferred to/from  
computer memory from/to a device or memory on the bus while the  
processor does something else. DMA is the fastest method of transferring  
data to/from computer memory.  
DNL  
Differential nonlinearity—a measure in least significant bit of the  
worst-case deviation of code widths from their ideal value of 1 LSB.  
DO  
Digital output.  
driver  
Software unique to the device or type of device, and includes the set of  
commands the device accepts.  
E
earth ground  
A direct electrical connection to the earth which provides a reference  
voltage level (called zero potential or ground potential) against which all  
other voltages in a system are established and measured. Also referred to as  
building ground.  
EEPROM  
ESD  
Electrically erasable programmable read–only memory–ROM that can be  
erased with an electrical signal and reprogrammed.  
Electrostatic Discharge—a high-voltage, low-current discharge of static  
electricity that can damage sensitive electronic components. Electrostatic  
discharge voltage can easily range from 1,000 to 10,000 V.  
F
F
Farad—a measurement unit of capacitance.  
FIFO  
First-in-first-out memory buffer—a data buffering technique that functions  
like a shift register where the oldest values (first in) come out first. Many  
DAQ products use FIFOs to buffer digital data from an A/D converter, or to  
buffer the data before or after bus transmission.  
floating signal  
sources  
Signal sources with voltage signals that are not connected to an absolute  
reference of system ground. Also called nonreferenced signal sources.  
Some common examples of floating signal sources are batteries,  
transformers, and thermocouples.  
© National Instruments Corporation  
G-5  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Glossary  
FPGA  
Field-programmable gate array.  
G
gain  
The factor by which a signal is amplified, often expressed in dB. Gain as a  
function of frequency is commonly referred to as the magnitude of the  
frequency response function.  
grounded signal  
sources  
Signal sources with voltage sources that are referenced to a system ground  
such as the earth or building ground. Also called referenced signal sources.  
H
h
Hour.  
Hertz.  
Hz  
I
I/O  
Input/output—the transfer of data to/from a computer system involving  
communications channels, operator interface devices, and/or data  
acquisition and control interfaces.  
in.  
Inches.  
INL  
Integral nonlinearity—for an ADC, deviation of codes of the actual transfer  
function from a straight line.  
IOH  
IOL  
Current, output high.  
Current, output low.  
Interrupt request.  
IRQ  
L
LED  
Light-emitting diode—a semiconductor light source.  
Least significant bit.  
LSB  
NI 6124/6154 User Manual  
G-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
M
m
Meter.  
master  
A functional part of a MXI/VME/VXIbus device that initiates data transfers  
on the backplane. A transfer can be either a read or a write.  
module  
A board assembly and its associated mechanical parts, front panel, optional  
shields, and so on. A module contains everything required to occupy one or  
more slots in a mainframe. SCXI and PXI devices are modules.  
MSB  
mux  
Most significant bit.  
Multiplexer—a switching device with multiple inputs that sequentially  
connects each of its inputs to its output, typically at high speeds, in order to  
measure several signals with a single analog input channel.  
N
NC  
Normally closed, or not connected.  
National Instruments.  
NI  
NI-DAQmx  
The latest NI-DAQ driver with new VIs, functions, and development tools  
for controlling measurement devices. The advantages of NI-DAQmx over  
earlier versions of NI-DAQ include the DAQ Assistant for configuring  
channels and measurement tasks for your device for use in LabVIEW,  
LabWindows/CVI, and Measurement Studio; increased performance such  
as faster single-point analog I/O; and a simpler API for creating DAQ  
applications using fewer functions and VIs than earlier versions of  
NI-DAQ.  
noise  
An undesirable electrical signal. Noise comes from external sources such  
as the AC power line, motors, generators, transformers, fluorescent lights,  
CRT displays, computers, electrical storms, welders, radio transmitters,  
and internal sources such as semiconductors, resistors, and capacitors.  
Noise corrupts signals you are trying to send or receive.  
P
PCI  
Peripheral Component Interconnect—a high-performance expansion bus  
architecture originally developed by Intel to replace ISA and EISA. It offers  
a theoretical maximum transfer rate of 132 Mbytes/s.  
© National Instruments Corporation  
G-7  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
pd  
Pull-down.  
PFI  
Programmable function interface.  
Programmable gain instrumentation amplifier.  
See channel.  
PGIA  
physical channel  
port  
1. A communications connection on a computer or a remote controller.  
2. A digital port consisting of four or eight lines of digital input and/or  
output.  
ppm  
Parts per million.  
Pull-up.  
pu  
PXI Express  
PCI Express eXtensions for Instrumentation—The PXI implementation  
of PCI Express, a scalable full-simplex serial bus standard that operates at  
2.5 Gbps and offers both asynchronous and isochronous data transfers.  
Q
quantization  
The process of converting an analog signal to a digital representation.  
Normally performed by an analog-to-digital converter (A/D converter or  
ADC).  
R
range  
The maximum and minimum parameters between which a sensor,  
instrument, or device operates with a specified set of characteristics.  
referenced signal  
sources  
Signal sources with voltage signals that are referenced to a system ground  
such as the earth or a building ground. Also called ground signal sources.  
rise time  
The time for a signal to transition from 10% to 90% of the maximum signal  
amplitude.  
rms  
Root mean square.  
RTSI bus  
Real-Time System Integration bus—the National Instruments timing bus  
that connects DAQ devices directly, by means of connectors on top of the  
devices, for precise synchronization of functions.  
NI 6124/6154 User Manual  
G-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
S
s
Seconds.  
Samples.  
S
S/s  
Samples per second—used to express the rate at which a digitizer or D/A  
converter or DAQ device samples an analog signal.  
scatter-gather  
SCXI  
The term used to describe very high-speed DMA burst-mode transfers that  
are made only by the bus master, and where noncontiguous blocks of  
memory are transparently mapped by the controller to appear as a seamless  
piece of memory.  
Signal Conditioning eXtensions for Instrumentation—the National  
Instruments product line for conditioning low-level signals within an  
external chassis near sensors so that only high-level signals are sent to DAQ  
devices in the noisy PC environment. SCXI is an open standard available  
for all vendors.  
sensor  
A device that responds to a physical stimulus (heat, light, sound, pressure,  
motion, flow, and so on) and produces a corresponding electrical signal.  
settling time  
The amount of time required for a voltage to reach its final value within  
specified limits.  
signal conditioning  
SOURCE  
The manipulation of signals to prepare them for digitizing.  
Source signal.  
system noise  
A measure of the amount of noise seen by an analog circuit or an ADC  
when the analog inputs are grounded.  
T
task  
NI-DAQmx—a collection of one or more channels, timing, and triggering  
and other properties that apply to the task itself. Conceptually, a task  
represents a measurement or generation you want to perform.  
terminal count  
The highest value of a counter.  
Gate hold time.  
tgh  
tgsu  
Gate setup time.  
© National Instruments Corporation  
G-9  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Glossary  
tgw  
Gate pulse width.  
THD  
Total harmonic distortion—the ratio of the total rms signal due to harmonic  
distortion to the overall rms signal, in dB or percent.  
THD+N  
thermocouple  
toff  
Signal-to-THD plus noise—the ratio in decibels of the overall rms signal to  
the rms signal of harmonic distortion, plus noise introduced.  
A temperature sensor created by joining two dissimilar metals. The  
junction produces a small voltage as a function of the temperature.  
An offset (delayed) pulse; the offset is t nanoseconds from the falling edge  
of the AI CONV CLK* signal.  
tout  
Output delay time.  
Period of a pulse train.  
See sensor.  
tp  
transducer  
tsc  
Source clock period.  
Source pulse width.  
tsp  
TTL  
Transistor-transistor logic—a digital circuit composed of bipolar  
transistors wired in a certain manner. A typical medium-speed digital  
technology. Nominal TTL logic levels are 0 and 5 V.  
V
V
Volts.  
VCC  
Nominal +5 V power supply provided by the PC motherboard.  
Common-mode noise and ground potential.  
Volts direct current.  
Vcm  
VDC  
VI, virtual instrument  
1. A combination of hardware and/or software elements, typically used  
with a PC, that has the functionality of a classic stand-alone instrument.  
2. A LabVIEW software module (VI), which consists of a front panel user  
interface and a block diagram program.  
VIH  
Volts, input high.  
NI 6124/6154 User Manual  
G-10  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Glossary  
VIL  
Volts, input low.  
Volts in.  
Vin  
Vm  
Measured voltage.  
Volts, output high.  
Volts, output low.  
Volts out.  
VOH  
VOL  
VOUT  
Vrms  
Volts, root mean square.  
Ground-referenced signal source.  
See channel.  
Vs  
virtual channel  
© National Instruments Corporation  
G-11  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
AI Convert Clock, 4-16  
Numerics  
10 MHz reference clock, 9-3  
100 kHz Timebase, 9-2  
20 MHz Timebase, 9-2  
80 MHz Timebase, 9-2  
AI Convert Clock Timebase, 4-17  
AI Reference Trigger, 4-19  
AI Sample Clock, 4-14  
AI Sample Clock Timebase, 4-16  
AI Start Trigger, 4-18  
connecting, 4-6  
terminal configuration, 4-2  
timing signals, 4-13  
A
A/D converter, 4-2  
timing summary, 4-13  
triggering, 4-6  
AC coupling connections, 4-9  
accessories  
field wiring considerations, 4-11  
I/O connector, 3-1  
circuitry, 5-2  
connecting signals, 5-4  
data generation methods, 5-2  
fundamentals, 5-1  
minimizing glitches, 5-2  
NI 6124, A-1  
NI 6154, A-7  
overview, 5-1  
triggering, 5-4  
NI 6124, A-4  
NI 6154, A-10  
acquisitions, digital waveform, 6-3  
AI Convert Clock signal, 4-16  
AI Convert Clock Timebase signal, 4-17  
AI Reference Trigger signal, 4-19  
AI Sample Clock signal, 4-14  
AI Sample Clock Timebase signal, 4-16  
AI Start Trigger signal, 4-18  
ai/ConvertClock, 4-16  
ai/ConvertClockTimebase, 4-17  
ai/ReferenceTrigger, 4-19  
ai/SampleClock, 4-14  
ai/SampleClockTimebase, 4-16  
ai/StartTrigger, 4-18  
analog output signals  
AO Pause Trigger, 5-10  
AO Sample Clock, 5-6  
AO Sample Clock Timebase, 5-8  
AO Start Trigger, 5-9  
analog triggering  
about, 11-2  
analog edge triggering, 11-3  
with hysteresis, 11-4  
analog window triggering, 11-6  
ANSI C documentation, xiv  
AO Pause Trigger signal, 5-10  
AO Sample Clock signal, 5-6  
AO Sample Clock Timebase signal, 5-8  
AO Start Trigger signal, 5-9  
analog input  
circuitry, 4-2  
data acquisition methods, 4-4  
fundamentals, 4-1  
overview, 4-1  
© National Instruments Corporation  
I-1  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Index  
applications  
routing, 9-1  
counter input, 7-2  
counter output, 7-19  
arm start trigger, 7-31  
common-mode  
noise  
differential ground-referenced  
signals, 4-9  
B
floating signals, 4-11  
differential signals, 4-9  
block diagram  
NI 6124, A-4  
NI 6154, A-10  
PFI input circuitry, 8-3  
PFI output circuitry, 8-3  
buffered  
signal range  
signal rejection considerations  
signals, 4-9  
period measurement, 7-7  
semi-period measurement, 7-8  
two-signal edge-separation  
measurement, 7-18  
bus  
connecting  
digital I/O signals, 6-9  
connecting signals  
analog input, 4-6  
digital I/O  
S Series isolated devices  
(NI 6154), 6-12  
(NI 6124), 6-9  
C
cables  
continuous pulse train generation, 7-21  
controlling counting direction, 7-2  
conventions used in the manual, xi  
counter  
NI 6124, A-4  
NI 6154, A-10  
calibration, 1-2, 2-4  
calibration certificate (NI resources), B-2  
cascading counters, 7-32  
Change Detection Event signal, 6-8  
channel  
input and output, 7-30  
output applications, 7-19  
terminals, default, 7-30  
Counter n A signal, 7-28  
Counter n Aux signal, 7-28  
Counter n B signal, 7-28  
Counter n Gate signal, 7-27  
Counter n HW Arm signal, 7-29  
Counter n Internal Output signal, 7-29  
Z behavior, 7-15  
choosing frequency measurement, 7-12  
clock  
10 MHz reference, 9-3  
external reference, 9-2  
generation, 9-1  
NI 6124/6154 User Manual  
I-2  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
Counter n Source signal, 7-26  
Counter n TC signal, 7-29  
Counter n Up_Down signal, 7-29  
Counter n Z signal, 7-28  
counter signals  
DAC FIFO, 5-2  
DACs, 5-2  
hardware for isolated devices, 2-3  
hardware for non-isolated devices, 2-2  
system overview, 2-1  
Counter n A, 7-28  
Counter n Aux, 7-28  
Counter n B, 7-28  
DAQ-PnP, 10-1  
Counter n Gate, 7-27  
Counter n HW Arm, 7-29  
Counter n Internal Output, 7-29  
Counter n Source, 7-26  
Counter n TC, 7-29  
DAQ-STC2, 2-3  
data generation methods, 5-2  
data transfer methods, changing, 10-2  
DC coupling  
connections, 4-9  
Declaration of Conformity (NI resources), B-2  
Counter n Up_Down, 7-29  
FREQ OUT, 7-30  
Frequency Output, 7-30  
counters, 7-1  
counter terminals, 7-30  
NI-DAQmx counter/timer pins, 7-30  
pins, 7-30  
cascading, 7-32  
connecting terminals, 7-30  
duplicate count prevention, 7-34  
edge counting, 7-2  
filters, 7-32  
generation, 7-19  
prescaling, 7-33  
single pulse generation with start  
trigger, 7-20  
device  
multiple synchronization, 9-3  
pinouts, 1-3  
specifications, 1-3  
DI change detection, 6-8  
di/SampleClock, 6-4  
DIFF mode  
about, 4-2  
connections for non-referenced or  
minimizing drift, 4-12  
differential connections  
for ground-referenced signal sources, 4-7,  
4-11  
synchronization modes, 7-37  
timing signals, 7-25  
triggering, 7-31  
for non-referenced or floating signal  
sources, 4-9  
counting edges, 7-2  
digital  
isolation (NI 6154), A-12  
© National Instruments Corporation  
I-3  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
triggering, 11-1  
waveform acquisition, 6-3  
waveform generation, 6-5  
digital I/O, 6-1  
edge counting, 7-2  
circuitry, 6-2  
buffered, 7-3  
on-demand, 7-2  
software, 6-10  
I/O protection, 6-7  
programmable power-up states, 6-7  
static DIO, 6-2  
edge-separation measurement  
NI-DAQmx, 7-37  
waveform acquisition, 6-3  
waveform triggering, 6-3  
digital I/O for isolated devices  
connecting signals, 6-9, 6-12  
I/O protection, 6-12  
X1, 7-14  
X2, 7-15  
X4, 7-15  
software, 6-13  
equivalent time sampling, 7-24  
exporting timing output signals using PFI  
terminals, 8-4  
static DIO, 6-11  
digital routing, 9-1  
digital signals  
Change Detection Event, 6-8  
connecting, 6-9  
DI Sample Clock, 6-4  
DO Sample Clock, 6-5  
digital waveform  
acquisition, 6-3  
generation, 6-5  
features  
counter, 7-32  
NI 6124, A-1  
NI 6154, A-7  
field wiring considerations, 4-11  
direct memory access (DMA), 10-2  
DMA, 10-2  
DO Sample Clock signal, 6-5  
do/SampleClock, 6-5  
documentation  
conventions used in manual, xi  
NI resources, B-1  
drivers (NI resources), B-1  
duplicate count prevention, 7-34  
enabling in NI-DAQmx, 7-37  
filter, 4-2  
filters  
counter, 7-32  
PFI, 8-6  
PXI_STAR, 9-9  
RTSI, 9-6  
NI 6124/6154 User Manual  
I-4  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
finite pulse train timing generation, 7-22  
FREQ OUT signal, 7-30  
frequency  
I/O protection, 6-7, 8-8  
input coupling, 4-2  
input polarity and range, 4-3  
input signals  
division, 7-24  
generation, 7-23  
generator, 7-23  
using PFI terminals as, 8-4  
using RTSI terminals as, 9-6  
installation  
hardware, 1-2  
NI-DAQmx, 1-1  
other software, 1-1  
G
instrument drivers (NI resources), B-1  
instrumentation amplifier, 4-2  
interrupt request (IRQ), 10-2  
IRQ, 10-2  
isolation (NI 6154), A-11  
digital isolation, A-12  
generations  
clock, 9-1  
continuous pulse train, 7-21  
digital waveform, 6-5  
frequency, 7-23  
pulse for ETS, 7-24  
pulse train, 7-21  
retriggerable single pulse, 7-20  
simple pulse, 7-19  
KnowledgeBase, B-1  
single pulse, 7-19  
single pulse with start trigger, 7-20  
glitches, 5-2  
L
LabVIEW documentation, xii  
M
H
Measurement Studio documentation, xiii  
measurements  
hardware, 1-2  
buffered period, 7-7  
buffered pulse-width, 7-5  
buffered semi-period, 7-8  
buffered two-signal edge-separation, 7-18  
choosing frequency, 7-12  
frequency, 7-9  
I
I/O connector  
NI 6124, A-2  
NI 6154, A-8  
period, 7-6  
signal descriptions  
isolated devices, 3-2  
non-isolated devices, 3-1  
position, 7-14  
pulse-width, 7-4  
semi-period, 7-7  
single period, 7-6  
© National Instruments Corporation  
I-5  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
single pulse-width, 7-4  
single semi-period, 7-8  
NI-DAQmx  
single two-signal edge-separation, 7-18  
documentation, xii  
measuring  
on-demand edge counting, 7-2  
other  
counters, 7-11  
low frequency with one counter, 7-9  
averaged, 7-9  
minimizing glitches on the output signal, 5-2  
MITE and DAQ-PnP, 10-1  
mux, 4-2  
pause trigger, 7-31  
N
PFI, 8-1, 8-2  
services, B-1  
exporting timing output signals using PFI  
.NET languages documentation, xiv  
NI 6124, A-4  
filters, 8-6  
analog output, A-1  
I/O protection, 8-8  
block diagram, A-4  
signals, 8-4  
cables and accessories, A-4  
features, A-1  
I/O connector pinout, A-2  
specifications, A-6  
PFI terminals as static digital I/Os, 8-5  
pinouts  
NI 6154, A-10  
block diagram, A-10  
digital I/O, 6-11  
counter default, 7-30  
device, 1-3  
NI 6124, A-2  
NI 6154, A-8  
RTSI connector, 9-4  
pins, default, 7-30  
polarity and reference selection, 5-2  
position measurement, 7-14  
buffered, 7-17  
isolation and digital isolators, A-11  
isolation barrier and digital isolators, 4-2,  
5-2  
specifications, A-13  
working voltage range, 4-4  
NI support and services, B-1  
power source, 3-3  
NI 6124/6154 User Manual  
I-6  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
power-up states, 6-7, 8-8, 8-9  
prescaling, 7-33  
reference  
clock  
programmable  
10 MHz, 9-3  
function interface (PFI), 8-1  
power-up states, 6-7, 8-8  
programmable function interface, 8-2  
programmable power-up states, 8-9  
programmed I/O, 10-2  
programming examples (NI resources), B-1  
pulse  
external, 9-2  
related documentation, xii  
retriggerable single pulse generation, 7-20  
routing  
clock, 9-1  
digital, 9-1  
RTSI, 9-4  
encoders, 7-16  
connector pinout, 9-4  
filters, 9-6  
using as outputs, 9-5  
signals, 9-6  
generation for ETS, 7-24  
train generation, 7-21  
continuous, 7-21  
pulse-width measurement, 7-4  
buffered, 7-5  
single, 7-4  
PXI  
S Series  
and PXI Express, 10-1  
clock, 10-1  
clock and trigger signals, 9-8  
trigger signals, 10-1  
triggers, 9-8  
devices, A-1  
specifications, xv  
sample clock  
edge counting, 7-3  
sample clock measurement, 7-17  
self-calibration, 1-2  
semi-period measurement, 7-7  
buffered, 7-8  
single, 7-8  
sensors, 2-5  
PXI Express  
and PXI, 10-1  
chassis compatibility, 10-1  
PXI_CLK10, 9-8  
PXI_STAR  
filters, 9-9  
signal conditioning, 2-5  
signal connections  
trigger, 9-8  
analog input, 4-6  
Q
signal descriptions  
quadrature encoders, 7-14  
S Series isolated devices (NI 6154), 3-2  
S Series non-isolated devices  
(NI 6124), 3-1  
R
real-time system integration bus, 9-4  
reciprocal frequency measurement, 7-11  
signal routing, RTSI bus, 9-4  
signal sources, 4-7  
© National Instruments Corporation  
I-7  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
signals  
software, 1-1  
AI Convert Clock, 4-16  
AI Convert Clock Timebase, 4-17  
AI Reference Trigger, 4-19  
AI Sample Clock, 4-14  
AI Sample Clock Timebase, 4-16  
AI Start Trigger, 4-18  
Change Detection Event, 6-8  
connecting digital I/O, 6-9  
connecting PFI input, 8-6  
Counter n A, 7-28  
AO applications, 5-11  
devices, 6-13  
NI resources, B-1  
programming devices, 2-6  
routing signals in, 9-10  
specifications  
device, 1-3  
NI 6124, A-6  
NI 6154, A-13  
Counter n Aux, 7-28  
start trigger, 7-31  
Counter n B, 7-28  
static DIO, 6-2  
Counter n Gate, 7-27  
S Series isolated devices (NI 6154), 6-11  
using PFI terminals as, 8-5  
support, technical, B-1  
synchronization modes, 7-37  
80 MHz source, 7-38  
Counter n HW Arm, 7-29  
Counter n Internal Output, 7-29  
Counter n Source, 7-26  
Counter n TC, 7-29  
Counter n Z, 7-28  
external source, 7-39  
other internal source, 7-39  
synchronizing multiple devices, 9-3  
synchronous counting mode, 7-34  
system timing controller  
counters, 7-25  
DO Sample Clock, 6-5  
terminals, 8-4  
FREQ OUT, 7-30  
Frequency Output, 7-30  
simple pulse generation, 7-19  
single  
terminals  
period measurement, 7-6  
point edge counting, 7-2  
retriggerable, 7-20  
connecting counter, 7-30  
Timebase  
100 kHz, 9-2  
20 MHz, 9-2  
80 MHz, 9-2  
with start trigger, 7-20  
pulse-width measurement, 7-4  
semi-period measurement, 7-8  
two-signal edge-separation  
timing output signals, exporting using PFI  
terminals, 8-4  
timing signals  
waveform generation, 5-6  
training, xv  
NI 6124/6154 User Manual  
I-8  
ni.com  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
training and certification (NI resources), B-1  
transducers, 2-5  
trigger  
generation  
digital, 6-5  
triggering, 6-3  
arm start, 7-31  
pause, 7-31  
PXI, 9-8  
waveform generation timing signals  
AO Pause Trigger, 5-10  
AO Sample Clock, 5-6  
AO Sample Clock Timebase, 5-8  
AO Start Trigger, 5-9  
summary, 5-6  
wiring, 4-11  
PXI_STAR, 9-8  
Star Trigger, 9-8  
start, 7-31  
triggering  
analog edge, 11-3  
analog edge with hysteresis, 11-4  
analog input, 4-6  
analog output, 5-4  
analog types, 11-3  
analog window, 11-6  
counter, 7-31  
X1 encoding, 7-14  
X2 encoding, 7-15  
X4 encoding, 7-15  
troubleshooting (NI resources), B-1  
two-signal edge-separation  
measurement, 7-17  
buffered, 7-18  
single, 7-18  
U
using  
PFI terminals  
as static digital I/Os, 8-5  
as timing input signals, 8-4  
to export timing output signals, 8-4  
RTSI  
as outputs, 9-5  
terminals as timing input signals, 9-6  
using PFI terminals  
as static digital I/Os, 8-5  
© National Instruments Corporation  
I-9  
NI 6124/6154 User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Miller Electric Welding System Big Blue 402P User Manual
Milwaukee Grinder 5211 User Manual
NEC Answering Machine 1093098 User Manual
New Holland Lawn Mower BB9O6O User Manual
NextBase Portable DVD Player SDV17 SQ User Manual
Omnimount Indoor Furnishings 10270 User Manual
Oster Blender 2609 User Manual
Palsonic Flat Panel Television TFTV435WS User Manual
Panasonic DVD Recorder DMR EZ47 User Manual
Parasound Stereo Amplifier Zpre2 User Manual