Leica Projection Television CM3050 S User Manual

Leica CM3050 S  
Cryostat  
Instruction Manual  
Leica CM3050 S – Cryostat  
English V1.2– 11/2006  
Always keep this manual near the instrument.  
Read carefully prior to operating the instrument.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Important note  
Serial No.: .............. ......................................................  
Year of Manufacture: ...............................................  
Country of Origin: Federal Republic of Germany  
The information contained in this document  
represents state-of-the-art technology as well as  
the current state of knowledge.  
Leica will not assume liability for errors that might  
be contained in this manual, nor for accidental  
damage or damage arising from the delivery,  
performance or use of this manual.  
Therefore, no claims can be made based on the  
text and illustrations in this instruction manual.  
Leica Microsystems Nussloch GmbH reserves the  
right to change technical specifications  
without notice since each of our products is  
subject to a policy of continuous improvement.  
This document is protected under copyright laws.  
Leica Microsystems Nussloch GmbH  
retains all rights related to this documentation.  
Any reproduction of text and illustrations - or any  
parts thereof - in form of printing, photocopying,  
microfiches, or other methods, including  
electronic systems, requires the prior written  
permission of Leica Microsystems Nussloch  
GmbH.  
The serial number and the year of manufacture  
are specified on the nameplate at the back of the  
instrument.  
© Leica Microsystems Nussloch GmbH  
Issued by:  
Leica Microsystems Nussloch GmbH  
Heidelberger Str. 17 - 19  
D-69226 Nussloch  
Germany  
Phone: +49 6224 143-0  
Fax: +49 6224 143-200  
Leica CM 3050 S – Cryostat  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
1.  
Important information................................................................................................................ 7  
1.1  
1.2  
Symbols used in this manual and .................................................................................................... 7  
Designated use ................................................................................................................................... 8  
1.3 Qualification of personnel .......................................................................................................... 8  
2.  
Sicherheit .................................................................................................................................... 9  
2.1  
2.2  
General information on instrument design and safe handling .................................................... 9  
Integrated safety devices ................................................................................................................. 9  
2.2.1 Locking the handwheel ................................................................................................................... 10  
2.2.2 Centering the handwheel grip ........................................................................................................ 10  
2.2.3 Emergency stop function (instruments with sectioningmotor only) ........................................ 11  
2.2.4 Knife guard ........................................................................................................................................ 11  
2.3  
Safety instructions: handling and operation of the instrument................................................. 12  
2.3.1 Transport ........................................................................................................................................... 12  
2.3.2 Site requirements ............................................................................................................................. 12  
2.3.3 Electrical connections ..................................................................................................................... 12  
2.3.4 Handling microtome knives/blades ............................................................................................... 13  
2.3.5 Knife guard/handwheel lock........................................................................................................... 13  
2.3.6 Motorized sectioning ....................................................................................................................... 13  
2.3.7 Defrosting/Handling frozen tissue ................................................................................................. 13  
2.3.8 Frozen parts of the instrument and frozen accessories ............................................................ 13  
2.3.9 Infectious/radioactive material ...................................................................................................... 14  
2.3.10 Disinfection and cleaning ............................................................................................................... 14  
2.3.11 Removing/reinstalling the microtome ........................................................................................... 15  
2.3.12 Display message ‘Dry microtome’ ................................................................................................. 15  
2.3.13 Maintenance .............................................................................................................................. 15  
3.  
Installation................................................................................................................................. 16  
3.1  
Site requirements ............................................................................................................................. 16  
3.1.1 General site requirements .............................................................................................................. 16  
3.1.2 Electrical connections ..................................................................................................................... 16  
3.2 Unpacking and installation ............................................................................................................. 16  
3.2.1 Repacking .......................................................................................................................................... 16  
3.3  
3.4  
Standard delivery ............................................................................................................................. 17  
Installing/inserting accessories..................................................................................................... 19  
3.4.1 Installing the handwheel ................................................................................................................. 19  
3.4.2 Inserting the accessories ............................................................................................................... 19  
3.4.3 The footswitch .................................................................................................................................. 20  
3.5 Prior to switching on the instrument ........................................................................................ 20  
Instruction Manual V1.2 - 11/2006  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
4.  
5.  
Instrument Properties .............................................................................................................. 21  
4.1 Overview ............................................................................................................................................ 21  
4.2 Technical Data ........................................................................................................................... 23  
Operation ................................................................................................................................... 25  
5.1  
5.2  
5.3  
Operating the instrument for the first time ................................................................................... 25  
Switching on/fuses........................................................................................................................... 26  
Control panel 1 .................................................................................................................................. 27  
5.3.1 Key functions in control panel 1..................................................................................................... 27  
5.3.2 Display functions in control panel 1 .............................................................................................. 27  
5.3.3 Display indications when switching on ........................................................................................ 28  
5.3.4 Menu functions: setting parameters of refrigeration, time and preset counter .................... 29  
5.3.5 Status display .................................................................................................................................... 32  
5.3.6 Indication of refrigeration state, section thickness setting and counters .............................. 33  
5.4  
Control panel 2 .................................................................................................................................. 34  
5.4.1 Selecting a sectioning mode / ........................................................................................................ 35  
5.4.2 Setting the sectioning window....................................................................................................... 36  
5.4.3 Selecting sectioning speed ............................................................................................................ 37  
5.4.4 Emergency stop ................................................................................................................................ 37  
5.4.5 Coarse feed keys .............................................................................................................................. 38  
5.4.6 Trimming and sectioning functions ............................................................................................... 39  
5.4.7 Setting the section thickness ......................................................................................................... 39  
5.5  
Daily operation of the instrument .................................................................................................. 40  
5.5.1 Freezing specimens onto specimen .............................................................................................. 40  
5.5.2 Freezing specimens onto specimen .............................................................................................. 41  
5.5.3 Optional accessories for freezing .................................................................................................. 41  
5.5.4 Inserting specimen discs into the .................................................................................................. 42  
5.5.5 Specimen orientation....................................................................................................................... 42  
5.5.6 Trimming / sectioning ...................................................................................................................... 43  
5.6  
Finishing work ................................................................................................................................... 44  
5.6.1 Daily routine steps............................................................................................................................ 44  
5.6.2 Switching off the instrument for a ................................................................................................. 44  
5.7  
Defrosting .......................................................................................................................................... 45  
5.7.1 Automatic chamber defrost cycle ................................................................................................. 45  
5.7.2 Manual chamber defrost cycle ...................................................................................................... 46  
5.7.3 Defrosting the specimen head ................................................................................................ 46  
Leica CM3050S – Cryostat  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of contents  
6.  
Troubleshooting, applications tips ....................................................................................... 47  
6.1  
6.1.1 Other error indications .................................................................................................................... 47  
6.2 Potential problems - causes and remedies ................................................................................. 48  
Displayed error messages .............................................................................................................. 47  
6.3 Reference chart for temperature settings (in minus °C) .................................................... 52  
7.  
Disinfection, cleaning and maintenance ............................................................................ 53  
7.1  
7.2  
7.3  
7.4  
7.5  
Safety instructions ........................................................................................................................... 53  
Spray disinfection with Leica Cryofect ......................................................................................... 53  
Disinfection with conventional ....................................................................................................... 54  
Cleaning ............................................................................................................................................. 54  
Removing / reinstalling the microtome ......................................................................................... 55  
7.5.1 Removing the microtome ................................................................................................................ 55  
7.5.2 Disassembling the specimen head................................................................................................ 55  
7.5.3 Reinstalling the microtome ............................................................................................................. 56  
7.6 Exchanging the fluorescent light lamp................................................................................... 57  
8.  
9.  
Warranty and service .............................................................................................................. 58  
Appendix .................................................................................................................................... 59  
1.  
Electrohydraulic height adjustment .............................................................................................. 59  
Ordering information - accessories ....................................................................................... 62  
2.  
10. EC Declaration of Conformity ..................................................................................... 63  
11. Decontamination Certificate (master) .................................................................................... 64  
Instruction Manual V1.2 - 11/2006  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. Important information  
The chapters of the Leica CM3050 S cryostat 1.1 Symbols used in this manual and  
their meaning  
instruction manual:  
Warnings  
appear in a grey box and are marked  
by a warning triangle  
Chapter 1 Structure of the manual:  
Table of contents  
.
Important information on this  
manual  
Chapter 2 Safety  
Read this chapter before  
operating the instrument!  
Chapter 3 Installation  
Notes  
i.e. important user information  
appears in a grey box and is marked  
by an information  
symbol.  
Unpacking and installation  
Standard delivery, information  
on accessories  
Figures in brackets refer to item  
numbers in drawings or to the  
drawings themselves.  
(5)  
(Fig.5)  
Chapter 4 Instrument properties  
Overview  
Technical data  
Instrument type:  
All information given in this instruction manual  
applies only to the instrument type indicated on  
the title page.  
Chapter 5 Operation  
Controls  
Setup, daily use and shut  
down  
A name plate, indicating the instrument serial  
number, is attached to the back of the instru-  
ment.  
Chapter 6 Trouble shooting  
and applications tips  
Operating errors  
Required in-  
formation for  
all inquiries:  
Display error messages  
Temperature reference chart  
For any inquir-  
please  
Chapter 7 Disinfection, cleaning  
and maintenance  
ies  
specify:  
Instrument  
type  
Chapter 8 Warranty and service  
Chapter 9 Electrohydraulic height adjustment  
Ordering information accessories  
EC Declaration of Conformity  
Serial-  
number  
Leica CM 3050 S – Cryostat  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. Important information  
General  
1.2 Designated use  
This instruction manual includes important  
instructions and information related to the  
operating safety and maintenance of the  
instrument.  
The Leica CM3050S is a powerful cryostat for  
routine as well as research applications in  
biology, medicine and industry.  
The instrument has been designed for rapid  
freezing and sectioning of tissue samples.  
The instruction manual is an important part of the  
product. It must be read carefully before using  
the instrument for the first time and must always  
be kept with the instrument.  
The instrument has not been designed for un-  
attended storage of tissue material.  
The instrument may only be operated within the  
scope of its designated use as described above  
and as per the instructions given in this manual.  
If additional requirements, which exceed the  
scope of this manual, are imposed by regula-tions  
and/or laws on accident prevention and  
environmental protection in the country of  
operation, appropriate instructions for  
compliance with such requirements must be  
added to this manual.  
Any other use of the instrument is considered  
improper.  
Read this instruction manual carefully before  
attempting to use or operate the instrument.  
1.3 Qualification of personnel  
The Leica CM3050S may only be operated by  
trained laboratory personnel.  
Please pay particular attention to  
(safety features, safety  
Please read this  
information, even if you are already  
familiar with the operation and use of  
other Leica products.  
All laboratory personnel designated to operate the  
instrument must carefully read the present  
instruction manual prior to starting work with the  
instrument.  
chapter  
instructions).  
2
-
Instruction Manual V1.2 - 11/2006  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Sicherheit  
2.1 General information on instrument  
design and safe handling  
2.2 Integrated safety devices  
The instrument is equipped with the following  
safety devices:  
This instrument has been built and tested in ac-  
cordance with the following safety regulations  
on electrical measuring, control, regulating and  
laboratory devices:  
Handwheel lock  
Handwheel grip centering  
Emergency stop function  
DIN EN 292  
DIN EN 61010-1  
EN 50082-1  
EN 55011  
(instruments with sectioning motor only)  
Knife holder equipped with knife guard  
The safety devices installed by the manufactu-  
rer of the instrument only constitute the basis of  
accident prevention.  
IEC 1000-4  
Mainly responsible for accident-free operation  
is above all the institution which owns the in-  
strument and, in addition, the designated per-  
sonnel who operate, service or repair the  
instrument.  
as well as according to the international quality  
standard  
DIN ISO 9001  
In order to maintain this condition and to ensure  
safe operation, the operator must observe the  
instructions and warnings contained in this in-  
struction manual.  
Leica CM3050 S – Cryostat  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Safety  
Instruments with sectioning motor:  
2.2.1 Locking the handwheel  
The sectioning motor is now blocked.  
Always cover the cutting edge with  
the knife guard and lock the hand-  
wheel:  
All instruments:  
Prior to doing any work on  
knife and/or specimen.  
Prior to exchanging specimens.  
During work breaks.  
The message ‘LOCKED’ in the display of control  
panel 1 indicates that the handwheel has been  
locked:  
C T - 3 0 ° C  
O T - 3 5 ° C  
L O C K E D  
The handwheel can be locked in 2 positions:  
with the grip in the uppermost position (left),  
with the grip in the lowest position (right).  
Unlocking:  
To unlock, push locking pin (2) to the left into  
position (2a).  
2
Display indication ‘LOCKED’ disappears.  
Instruments with sectioning motor:  
The sectioning motor can now be activated  
again.  
2.2.2 Centering the handwheel grip  
2
During motorized sectioning, for  
safety reasons always center the  
handwheel grip!  
Locking:  
Rotate handwheel, until grip (1) is in upper or  
lower position.  
To lock, press pin (2) to the right into position  
(2b).  
To center grip (1), pull outwards and pivot  
into center of handwheel.  
When released, the grip locks into position.  
--> The upper locking position for pin (2) is  
marked by a black dot (4).  
Instruction Manual V1.2 - 11/2006  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.  
Safety  
2.2.3 Emergency stop function  
(instruments with sectioning  
motor only)  
2.2.4 Knife guard  
All knife holders are equipped with a knife  
guard (--> see separate instruction manuals on  
knife holders).  
The emergency stop is activated via the red  
emergency stop button in control panel 2 or via  
the footswitch  
Always cover the cutting edge with  
the knife guard:  
• Prior to doing any work on knife  
and/ or specimen.  
• Prior to exchanging specimens.  
• During work breaks.  
Control panel 2  
Footswitch  
Activating the emergency stop function  
Press emergency stop button or step on  
footswitch forcefully.  
-->  
STOP (red) lights up.  
As soon as the emergency stop function is acti-  
vated, the sectioning motor stops.  
Deactivating the emergency stop  
To deactivate, rotate emergency stop button  
in direction of arrow.  
If the emergency stop function has  
been activated by the footswitch,  
unlocking is not necessary (func-tion  
is unlocked as soon as the footswitch  
is released).  
Leica CM3050 S – Cryostat  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Safety  
2.3 Safety instructions: handling and operation of the instrument  
2.3.3 Electrical connections  
2.3.1 Transport  
Do not use extension cords for connecting  
the instrument to mains.  
-->Fire hazard!  
To avoid severe damage to the instrument by  
running it while the compressor oil is dis-  
placed from its regular position:  
-->Instrument malfunctions caused by  
voltage drop.  
Do not tilt the instrument,  
only transport in an upright position.  
After transport, wait at least 4 hours be-  
fore turning the instrument on.  
During the start-up phase of the compressor,  
the nominal voltage must not drop below the  
values specified in chapter 4.2 ‘Technical  
data’!  
-->The compressor needs a start-up current  
at between 25 and 35 A (see chapter 4.2  
‘Technical data’)  
-->The compressor oil may have been  
displaced during transport and must  
settle to its original position  
be-  
fore switchingthe instrument on.  
Otherwise, the instrument may be  
severely damaged.  
Ensure uniform current supply according to  
specifications.  
2.3.2 Site requirements  
-->Electrical power supply deviating from  
specifications damages the instrument.  
Do not operate the instrument in rooms with  
explosion hazard.  
Therefore, arrange for the electrical in-  
stallations on site to be checked by a  
trained professional and make sure any  
necessary upgrades are installed!  
Have the circuit protected by a fuse of its  
own!  
To ensure proper instrument function main-  
tain a minimum distance of 10 cm between  
walls and/or furniture and all sides of the in-  
strument!  
Do not connect any other consumers to  
the same circuit.  
Prior to connecting the instrument to  
mains, make sure the electrical power  
supply in your laboratory corresponds to  
the values indicated on the instrument  
nameplate (located at the rear of the in-  
strument)!  
Instruction Manual V1.2 - 11/2006  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Safety  
2.3.4 Handling microtome knives/blades  
2.3.7 Defrosting/Handling frozen tissue  
Danger!  
-->Microtome knives and disposable blades  
have extremely sharp cutting edges and  
can cause serious injuries. Therefore:  
Never leave specimens unattended in the  
cryochamber over an extended period of  
time!  
Handle knives / blades with utmost care.  
Never leave any knives / blades in unpro-  
tected places!  
Never place a knife, no matter where,  
with the cutting edge facing upwards.  
Never try to catch a falling knife.  
Always insert the specimen before  
inserting the knife.  
-->In case of power failure or instrument  
failure, or during the automatic defrost cy-  
cle, tissue material can be destroyed.  
During the defrost cycle the cryochamber is  
partially warmed. - Therefore:  
Remove sensitive specimens from the  
chamber prior to defrosting.  
If automatic defrosting is programmed to  
take place during the night, remember to  
remove all specimens from the cryo-  
chamber prior to leaving work.  
2.3.5 Knife guard/handwheel lock  
Always cover the cutting edge with the knife  
guard and lock the handwheel:  
2.3.8 Frozen parts of the instrument and  
frozen accessories  
Prior to doing any work on knife and / or  
specimen.  
Prior to exchanging specimens.  
During work breaks.  
Prolonged contact of bare skin to frozen  
surfaces of the instrument or to frozen  
accessories (specimen discs, knife holder,  
shelves etc.) can cause frostbite.  
2.3.6 Motorized sectioning  
Wear protective gloves.  
Do not interrupt sectioning / trimming by set-  
ting the sliding potentiometer to zero speed!  
-->This does not really switch the sectioning  
function off - it only operates at ‘0’ speed.  
If the sliding potentiometer is accidental-  
ly moved, the instrument will resume sec-  
tioning immediately (risk of injury)!  
During motorized sectioning, always center  
the handwheel grip!  
Leica CM3050 S – Cryostat  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Safety  
2.3.9 Infectious/radioactive material  
2.3.10 Disinfection and cleaning  
Prior to disinfection, switch the instrument  
off and unplug it from mains.  
Danger!  
Use caution when working with potentially  
infectious specimens:  
-->Risk of infection!  
For removal of the microtome from the cryo-  
chamber see chapter 2.3.11 ‘ Removing the  
microtome’.  
When working with potentially infectious /  
radioactive specimens:  
For disinfection, wear protective gear:  
(gloves, mask, lab coat etc.)!  
Wear protective clothes (gloves, protec-  
tive boots, mask, lab coat), in compliance  
with radiation safety regulations and/or  
in-house regulations on handling infec-  
tious / radioactive material.  
For disinfection, only use alcohol-based dis-  
infectants!  
Do not use solvents (xylene, acetone etc.)  
for cleaning or disinfection!  
When working with radioactive specimens:  
Do not spray disinfectants into the evapora-  
tor!  
-->Risk of icing!  
Comply with applicable radiation safety  
regulations!  
Dispose of radioactive specimen waste ac-  
cording to applicable regulations.  
Explosion hazard when working with alcohol:  
Make sure the premises are appropriately  
ventilated!  
When using disinfectants and detergents,  
comply with all safety instructions supplied  
by the manufacturer of the product!  
Dispose of waste liquids from disinfection/  
cleaning as well as of sectioning waste ac-  
cording to applicable regulations on disposal  
of special category waste!  
Disinfected accessories must be thoroughly  
dry when reinserting them into the chamber!  
-->Risk of icing!  
Make sure the chamber is completely dry  
before switching the instrument back on:  
-->Explosion hazard through alcohol vapors!  
-->Risk of icing!  
Instruction Manual V1.2 - 11/2006  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Safety  
2.3.11 Removing/reinstalling the  
microtome  
2.3.12 Display message ‘Dry microtome’  
If the error message ‘Dry Microtome’ is dis-  
played in control panel 1, the following has  
happened:  
Before removing the microtome:  
Switch instrument off.  
Cryochamber refrigeration has been in-  
terrupted for an extended period of time  
(e.g. power failure), causing the chamber  
temperature to rise into the positive  
digits.  
Unplug from mains.  
Place handwheel grip in lowest position and  
lock.  
-->When removing the microtome, the  
specimen head must always be lok-  
ked in the lowest position.  
Otherwise the upper part of the slot  
cover might be bent and consequently  
damaged!  
• If this message appears, do not switch on  
the instrument but remove the microtome  
from the chamber, disinfect, if necessary,  
and dry thoroughly before reinstalling it  
into the chamber (see chapter 7.1 to 7.5).  
When removing the microtome:  
2.3.13 Maintenance  
Wear gloves when removing the micro-  
tome while it is still frozen.  
-->Risk of frost bite!  
On instruments with specimen cooling:  
do not distort the refrigerating tube!  
If distorted it might break, causing ex-  
Only technical service engineers authorized  
by Leica may access the internal compo-  
nents of the instrument for service and re-  
pair.  
The fluorescent light lamp (chamber illumi-  
nation), unless broken or splintered, can be  
replaced by the user:  
tremely cold refrigerant to escape.  
> Risk of frost bite!  
- -  
Before reinstalling the microtome:  
Switch off mains switch!  
Microtome must be completely dry.  
>Humidity in the interior of the micro-  
tome freezes and causes  
--  
Unplug the instrument from mains!  
If the lamp is broken or splintered:  
Have lamp replaced by Leica Technical  
Service!  
microtome malfunctions and/or  
damage to the microtome.  
-->Risk of injury!  
Use only those replacement lamps that cor-  
respond to technical specification (see  
chapter 4.2 ‘Technical Data’).  
All accessories/tools removed from the  
cryochamber must be thoroughly dry be-  
fore putting them back into the chamber!  
-->Risk of icing!  
Leica CM3050 S – Cryostat  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Installation  
3.1 Site requirements  
3.2 Unpacking and installation  
Make sure to read and follow all  
safety instructions in chapter 2.3.2  
‘Site requirements’!  
Unpacking instructions are always located in a  
transparent protective envelope on the outside  
of the instrument shipping crate.  
3.1.1 General site requirements  
Make sure to read and follow all  
safety instructions provided in  
chapter 2.3.1 ‘Transport’ and on the  
unpacking instructions!  
No direct sunlight.  
Electrical power supply within distance  
(length of power cord = approx. 4 meters -  
do not use extension cords!).  
3.2.1 Repacking  
No draft (caused by air conditioning etc.).  
Even floor surface.  
We recommend to keep the original shipping  
crate and the unpacking instructions for the  
Leica CM3050 S.  
Practically vibration-free floor.  
Handwheel easily accessible.  
For repacking, proceed as per unpacking in-  
structions, in reverse order.  
Room temperature constantly below +22 °C.  
Relative humidity of air maximum 60 %.  
Available models  
High ambient temperature and/or high  
air humidity negatively affect  
instrument cooling performance!  
Basic instrument  
- with sectioning motor  
- without specimen cooling  
3.1.2 Electrical connections  
Basic instrument  
- without sectioning motor  
- with specimen cooling  
Make sure to read and follow all  
safety instructions in chapter 2.3.3  
‘Electrical connections’!  
Basic instrument  
- with sectioning motor  
- with specimen cooling  
Instruction Manual V1.2 - 11/2006  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Installation  
3.3 Standard delivery  
1
1
1
1
1
Cryostat with microtome ............................................................................. 0460 32037  
Handwheel with marking, antibacterial ...................................................0471 42558  
Heat extractor, stationary...........................................................................0369 11197  
Low-temperature stabilizer for heatextractor ........................................0452 27919  
90° prism for direct specimen freezing on specimen head  
with clamping screw to facilitate specimen trimming .......................... 0443 25949  
Set of specimen discs .................................................................................0470 43550  
- 4 Specimen discs, 25 mm .........................................................................0416 19275  
- 4 Specimen discs, 30 mm .........................................................................0370 08587  
Storage shelf, right ...................................................................................... 0443 25723  
Storage shelf, left ......................................................................................... 0443 33549  
Section waste tray .......................................................................................0400 26817  
Foot switch with protective guard ............................................................0502 29977  
Rubber mat .................................................................................................... 0443 25732  
Freezing shelf cover ....................................................................................0443 30783  
Brush shelf .................................................................................................... 0398 13088  
Tool set........................................................................................................... 0436 43463  
-1 Brush, fine ................................................................................................. 0183 28642  
-1 Leica-brush ............................................................................................... 0183 30751  
-1 Allen key, size 1.5 .....................................................................................0222 10050  
-1 Allen key, size 2.5 .....................................................................................0222 04137  
-1 Allen key, size 3.0 .....................................................................................0222 04138  
-1 Allen key, size 4.0 .....................................................................................0222 04139  
-1 Allen key with spherical head, size 4.0 ................................................0222 32131  
-1 Allen key, size 5.0 .....................................................................................0222 04140  
-1 Allen key with handle, size 5.0 ...............................................................0194 04760  
-1 Allen key, size 6.0 .....................................................................................0222 04141  
-1 Single-head wrench, size 13..................................................................0330 33149  
-1 Single-head wrench, size 16..................................................................0330 18595  
Bottle of OCT-Compound, mounting medium  
1
1
1
1
1
1
1
1
1
1
for cryosectioning, 125 ml...........................................................................0201 08926  
Bottle of cryostat oil, type 407, 50 ml ........................................................0336 06098  
Pair of cut resistant gloves, size S ............................................................0340 40859  
Instruction manual Leica CM3050 S - G/E/F/S ........................................0708 37109  
1
1
1
In addition to the above:  
Instruments with specimen cooling:  
1
Configured instruments:  
Instruments with sectioning motor:  
1
90° Prism  
for low temperature quick freezing  
1
1
Knife holder base  
Knife holder with accessories  
Footswitch with  
protective guard  
Leica CM 3050 S – Cryostat  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.  
Installation  
In addition to the above:  
Instruments with specimen cooling:  
90° Prism for low temperature quick freezing  
Configured instruments:  
1
1
1
Knife holder base  
Knife holder with accessories  
Instruments with sectioning motor:  
Footswitch with protective guard  
1
Further accessories:  
Further accessories which you ordered will be included in the box containing  
the standard delivery items.  
Knife holders are delivered with anti-roll guide, knife guard, and a  
separate instruction manual.  
In case of non-configured instruments, the knife holder is not a  
part of standard delivery but must be ordered separately.  
Check all delivered parts against the packing list and against your order  
to verify whether the delivery is complete!  
If there is any difference, contact your local Leica office immediately.  
Instruction Manual V1.2 - 11/2006  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Installation  
3.4 Installing/inserting accessories  
3.4.1 Installing the handwheel  
3.4.2 Inserting the accessories  
Place the rubber mat on top of the housing.  
Insert the storage shelves into the cryo-  
chamber.  
Insert pin (1) of the handwheel shaft into  
hole (2).  
Install the stationary heat extractor into the  
quick-freeze shelf (see also drawing in  
chapter 5.5.1 ‘Freezing specimens onto  
specimen discs’).  
Place spring washer (3) onto screw (4) as  
shown.  
Tighten screw (4) with an Allen key.  
Cover with selfadhesive disc (5).  
Insert the low temperature stabilizer into the  
quick freeze shelf (it must be located in the  
pivoting range of the heat extractor - see  
also drawing in chapter 5.5.1 ‘Freezing spec-  
imens onto specimen discs’).  
Insert section waste tray and brush shelf.  
Install knife holder base onto microtome  
base plate and clamp.  
Install knife holder and clamp (see knife  
holder instruction manual for details).  
Place knife case with knife into chamber to  
precool.  
Place all tools needed for section prepara-  
tion into the chamber.  
For purposes of transport (e.g. narrow  
doors), the handwheel can be  
removed.  
Close the sliding window.  
For  
individual parts:  
--> see Chapter 4.1 ‘Overview’.  
a
complete overview of all  
To remove the handwheel, proceed as de-  
scribed above but in reverse order.  
Leica CM 3050 S – Cryostat  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Installation  
3.4.3 The footswitch  
Connecting the footswitch  
Insert footswitch into port (1) and secure.  
1
Function  
The footswitch performs the same task as the  
RUN/STOP and RUN/ENABLE keys (activating/  
deactivating motorized sectioning / trimming).  
In addition, the footswitch can be used to acti-  
vate the emergency stop function.  
3.5 Prior to switching on the instrument  
After transport, observe a waiting  
period of at least 4 hours before  
turning the instrument on! - See also  
safety instructions in chapter 2.3.1  
‘Transport’.  
Models with footswitch:  
Have you observed all safety  
instructions in chapters 2.3.2 ‘Site  
requirements’ and 2.3.3 ‘Electrical  
connections’?  
All instruments with sectioning motor.  
If not:  
--> Please read chapters 2.3.2 and  
2.3.3!  
Important note:  
In all instrument models that are  
delivered with footswitch, the  
footswitch must be installed!  
-
Otherwise the instruments are not  
functional.  
Insert mains plug into wall outlet.  
Continue with chapter 5.1 ‘Setting up the in-  
strument’.  
Instruction Manual V1.2 - 11/2006  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Instrument Properties  
4.1 Overview  
1
2
3
4
5
6
7
Cryostat CM3050S  
Control panel 1  
Control panel 2  
Storage shelf, left  
Storage shelf, right  
Rubber mat  
4
2
5
3
Mains switch /  
Automatic cutout  
for instrument,  
Automatic cutout  
for sectioning motor,  
Footswitch port  
Quick-freeze shelf  
Stationary  
6
8
9
heat extractor  
10 Mobile heat extractor  
11 Specimen disc  
12 Thermoblock (optional)  
13 Section waste tray  
14 Brush shelf  
9
10  
15 Specimen head w/o  
specimen cooling  
16 Specimen head with  
specimen cooling  
(Option)  
17 90° Prism  
1
(instruments with  
specimen cooling  
only)  
11  
8
7
18 Knife holder base  
19 Knife holder CE  
20 Knife holder CN  
21 Knife holder CS  
22 Footswitch with  
protective guard  
12  
22  
Leica CM3050S – Cryostat  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Instrument Properties  
15  
16  
11  
11  
13  
17  
21  
20  
19  
18  
Instruction Manual V1.2 - 11/2006  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Instrument properties  
4.2 Technical Data  
Operating temperature range (ambient temperature): + 18°C to + 40°C. All specifications related to tempe-  
rature are valid only up to an ambient temperature of + 22°C and relative air humidity of less than 60%!  
Type  
CM3050S-10  
CM3050S-1  
CM3050S-8  
CM3050S-9  
Marks of conformity  
Nominal voltage  
Nominal frequency  
Power draw  
-
-
-
-
100 V AC 10% 230 V AC 10% 240 V AC 10% 100 V AC 10%  
50 Hz  
1800 VA  
50 Hz  
1800 VA  
25 A eff.  
I
50 Hz  
1800 VA  
25 A eff.  
I
60 Hz  
1800 VA  
30 A eff.  
I
Max. start-up current for 5 sec 35 A eff.  
I
Protective fuse  
Automatic cutout  
T15A M3  
2
T10A T1  
2
T10A T1  
2
T15A M3  
2
Pollution degree  
Overvoltage installation category II  
II  
1800 J/s  
II  
1800 J/s  
II  
1800 J/s  
Heat emission (max.)  
1800 J/s  
Refrigeration system  
50 Hz  
60 Hz  
Cryochamber  
Temperature selection range  
Defrosting  
0° C to -40°C (+ 2 K / - 0 K)  
Automatic hot-gas defrost cycle,  
0° C to -40°C (+2 K / - 0 K)  
Automatic hot-gas defrost cycle,  
programmable in 15-min. increments;  
1 automatic defrost cycle per 24 hours.  
Duration: 6 - 12 minutes  
Manual defrost cycle  
programmable in 15-min increments;  
1 automatic defrost cycle per 24 hours.  
Duration: 6 - 12 minutes  
Manual defrost cycle  
690 W  
Refrigeration capacity  
Safety factor  
Refrigerant  
690 W  
3
3
280 g 5g refrigerant R 404A *  
280 g 5g refrigerant R 404A*  
Compressor oil  
0.6 l EMKARATE RL-22S, ICI *  
0.6 l EMKARATE RL-22S, ICI *  
Quick-freeze shelf  
Maximum temperature  
Number of quick-freeze stations 10  
- 43°C (+ 0 K / - 2 K)  
- 43°C (+ 0 K / - 2 K)  
10  
Specimen cooling (Option)  
Temperature range  
Defrosting  
Refrigeration capacityꢀ  
Safety factor  
-10° C to -50°C 3K  
Manual defrost cycle (electric)  
320 W  
-10° C to -50°C 3K  
Manual defrost cycle (electric)  
320 W  
3
3
Refrigerant  
Compressor oil  
185 g 5g refrigerant R 404A *  
0.4 l alpha 22, Kyodo *  
or RENISO E22, Fuchs *  
185 g 5g refrigerant R 404A *  
0.4 l alpha 22, Kyodo *  
or RENISO E22, Fuchs *  
Leica CM3050S – Cryostat  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Instrument properties  
*) Refrigerant and compressor oil to be replaced only by trained and authorized service  
personnel!  
CM3050S-3  
c-UL  
120 V AC 10% 208 V AC 10% 230 V AC 10%  
CM3050S-6  
-
CM3050S-7  
-
Microtome  
Rotary microtome  
Section thickness setting  
Overall specimen feed  
Vertical stroke  
0.5 - 300 μm  
25 mm  
59 mm  
60 Hz  
1800 VA  
35 A eff.  
I
60 Hz  
1800 VA  
30 A eff.  
I
60 Hz  
1800 VA  
25 A eff.  
I
Specimen retraction  
Maximum specimen size  
Specimen orientation  
50 μm  
40 x 55 mm  
8° (x-, y-, z-axis)  
T15A T1  
2
T12A T1  
2
T10A T1  
2
Trimming (6 thickness settings)  
Motorized coarse feed  
5 - 150 μm  
II  
1800 J/s  
II  
1800 J/s  
II  
1800 J/s  
- slow  
- fast  
500 μm/s  
1,000 μm/s  
Sectioning motor (Option):  
Sectioning speed  
- slow  
Fluorescent light lamp (cryochamber illumination):  
min. 0.1 mm/s  
max. 100 mm/s  
50 Hz version:  
60 Hz version:  
Osram Dulux S 11 W/21  
Light color ‘LUMILUX hellweiß’  
v
min.  
210 mm/s  
0.1 mm/s  
(brilliant white)  
max  
- fast  
Osram Dulux S 13 W/21  
Light color ‘LUMILUX hellweiß’  
max. 170 mm/s  
210 mm/s  
v
(brilliant white)  
max  
Cryocabinet  
Dimensions  
according to IEC-1010, UL 3101  
1
2
Width (w/o handwheel)  
Width (including handwheel)  
Depth  
800 mm  
882 mm  
766 mm  
840 mm  
1040 mm  
according to CECOMAF:  
Liquid temperature 45°C  
Evaporation temperature -25°C  
Height (arm rest level)  
Overall height  
Weight  
(incl. microtome, w/o specimen cooling)  
ca. 180 kg  
Instruction Manual V1.2 - 11/2006  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.1 Operating the instrument for the first time  
Prior to setting up the instrument, make sure to read the instruction manual  
carefully and in full.  
When setting up the instrument, please follow the individual chapters in the  
following order:  
First, study chapters 1, 2, and 3 - install the instrument as described in  
chapter 3.  
The overview in chapter 4.1 is helpful during the unpacking and instal-  
lation procedure in order to identify individual parts and place them in  
the correct location.  
Go on to chapter 5.  
Read chapters 5.1, 5.2, 5.3 and 5.4 to learn all about the controls.  
Next, start to work with the instrument (chapter 5.5). Follow the instruc-  
tions in chapter 5.5.  
Chapter 5.5.2 applies to instruments with specimen cooling only.  
The temperature chart in chapter 6.3 is useful when it comes to select-  
ing appropriate sectioning temperatures for different types of speci-  
mens.  
When disinfecting / cleaning the instrument and/or removing / rein-  
stalling the microtome, please follow the instructions in chapter 7.  
In case of problems and malfunctions, please refer to chapters 6.1 and  
6.2. These chapters list the most common types of problems which  
many times have simple causes that can be eliminated by the actual  
user.  
If you find ‘Call Technical Service’ in the ‘Remedy’ column, please do  
so. Never try to solve these problems yourself!  
Leica CM3050 S – Cryostat  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Automatic cutout for sectioning motor:  
5.2 Switching on/fuses  
The instrument is equipped with an automatic  
cutout (2) for the sectioning motor.  
If the sectioning motor is overloaded (e.g. when  
sectioning extremely hard specimens etc.) the  
automatic cutout (2) will be triggered.  
To make the instrument functional again after  
the automatic cutout (2) has been triggered:  
2
(Automatic cutout  
for sectioning  
motor)  
Push automatic cutout (2) inwards until it  
locks into position.  
1
(Combination Mains switch /  
automatic cutout for the  
instrument)  
Switching on  
To switch on the instrument, place switch (1)  
in upper lock-in position.  
Mains switch / automatic cutout for the  
instrument:  
In addition to its function as ON/OFF switch,  
switch (1) is also the automatic cutout for the in-  
strument.  
If the automatic cutout has been triggered (e.g.  
caused by overcurrent):  
Put switch (1) back in upper lock-in position.  
Instruction Manual V1.2 - 11/2006  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.3 Control panel 1  
5.3.2 Display functions in control panel 1  
5.3.1 Key functions in control panel 1  
The display in control panel 1 combines several  
different functions:  
ON/OFF key for cryochamber illumination.  
1. When switching the instrument on:  
To select the menu items for setting /  
changing instrument parameters.  
A series of standard indications is displayed,  
during which two basic parameters can be se-  
lected (with the arrow keys):  
Language (of display indications) and speed  
level of sectioning motor.  
See chapter 5.3.3 ‘Display indications when  
switching on’.  
In the status indication, pressing the ar-  
row keys displays the individual current  
parameters.  
In the menu, pressing the arrow keys  
changes set values.  
2. Menu function (activated via menu key):  
Via the menu key a number of parameters can  
be selected, then be activated /deactivated via  
the lock/unlock key and/or changed via the ar-  
row keys.  
As soon as a value is displayed, it is au-  
tomatically stored.  
Activates the manual defrost cycle (cry-  
ochamber). This key is functional only  
when one of the menu items ‘SET TEMP  
CT ... °C’ or ‘SET TEMP OT ... °C’ (last  
one only applies to instruments with  
specimen cooling) has been selected  
first.  
See chapter 5.3.4 ’Menu functions: Setting param-  
eters of refrigeration, time and preset counter’.  
3. Status indication function:  
The actual status of selected parameters is in-  
dicated in the display screen.  
Some of those parameters are displayed auto-  
matically (e.g. sectioning stroke phase, coarse  
feed position, handwheel locked).  
Others can be displayed by pressing the arrow  
keys; e.g. selected section thickness, preset  
counter status etc.  
See chapter 5.3.5 ‘Status indication’ and chap-  
ter 5.3.6 ‘Displaying state of refrigeration, sec-  
tion thickness and counters’.  
Lock / unlock key:  
- Locks the display to protect against  
accidental changes of set parameters.  
- Activates / deactivates a number of  
functions.  
Leica CM 3050 S – Cryostat  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.3.3 Display indications when switching on  
Meaning of indication:  
Operator activity:  
---  
Initialization phase 1  
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8  
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8  
Initialization phase 2  
---  
C M 3 0 5 0 S  
V s . X X . X X  
Software version  
---  
E N G L I S H  
Selected display language  
Select language via  
keys  
C U T T I N G M O T O R  
S L O W  
Sectioning motor speed  
level ‘slow’ / (fast)  
Select speed level via  
keys  
Temperature prior to  
switching off < 0 °C  
yes  
no  
Temperature upon  
switching on 0 °C  
no  
C T . . . . ° C  
P O W E R  
O T .  
.
. . ° C * )  
F A I L U R E  
yes  
Acknowledge indication by  
pressing once briefly.  
C T . . . . ° C  
O T .  
.
. . ° C * )  
C T . . . . ° C  
O T .  
.
. . ° C * )  
D R Y M I C R O T O M E  
1 2 : 0 0  
Standard indication ‘locked’ is displayed:  
(Display background dark)  
If this indication is displayed: Do not switch on  
the instrument!  
To unlock: press  
ground lights up:  
until the display back-  
Prior to switching on, both microtome and  
accessories must be completely dry!  
Remove microtome/accessories, disinfect if  
applicable, then dry thoroughly and reinstall.  
(--> See chapters 7.1 - 7.5).  
C T . . . . ° C  
O T .  
.
. . ° C  
H O M E  
❄ ❄  
*) Indication in instruments with specimen cooling. Instruments without specimen cooling: only CT (centered) is displayed.  
Instruction Manual V1.2 - 11/2006  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.3.4 Menu functions: setting parameters of refrigeration, time and preset counter  
Once the instrument has gone through the initial  
sequence of indications, you can go on to the  
menu functions to select and/or change instru-  
ment parameters.  
1. Specimen temperature:  
(instruments with specimen cooling only)  
Press key until menu item ‘Set specimen  
temperature’ is displayed:  
If display is locked:  
Press key until display background  
lights up.  
S E T  
T E M P O T - - - ° C  
- -  
Press key briefly.  
The line in the specimen temperature  
field is replaced by an asterisk.  
Next, press the menu key. With each key stroke  
the following parameters will be displayed (dis-  
play sequence as listed):  
S E T  
T E M P O T - - - ° C  
- ❄  
1. Specimen temperature  
(in instruments with specimen cooling only)  
(-10 - -50 °C)  
From this point on the refrigeration sys-  
tem is activated and the desired set  
temperature can be selected with one  
of the arrow keys.  
2. Cryochamber temperature  
(0 - -40 °C)  
Press arrow key and release upon  
reaching the desired value.  
3. Selected number of sections - preset  
counter  
(ON/OFF and setting selection from 1 to 999)  
C T . . . . ° C  
H O M E  
O T - 3 5 ° C  
- ❄  
4. Maximum specimen temperature  
(ON/OFF)  
Note on the arrow key function:  
Each time one of the arrow keys is pressed  
briefly, the value displayed in the menu will be  
changed one step up or down.  
If an arrow key is pressed and held, the value  
will change continually (auto-repeat function).  
When holding the key, scrolling speed will in-  
crease after a while.  
5. Real time  
(in 1-minute steps)  
6. Defrost time (starting time of automatic  
cryochamber defrosting)  
(in 15-minute steps)  
7. Duration of defrost cycle  
(cryochamber defrosting)  
(6 to 12 minutes duration)  
As soon as a value is displayed, it is at the same  
time automatically stored.  
Menu items 5 and 6 displayed on 24 hour clock.  
Leica CM 3050 S – Cryostat  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Now there are two options:  
1. Increase or decrease the actual reference  
value by pressing or  
2. Press and simultaneously in order to  
2. Cryochamber temperature:  
Press key until menu item ‘Set chamber  
temperature’ is displayed:  
.
--> Continue as described for  
specimen temperature.  
reload the former reference value (‘0’ when  
setting up the instrument) and then press ei-  
3. Preset counter  
(instruments with sectioning motor only):  
ther  
or  
individually in order to select  
the desired reference value. Depending on  
where you start from and where you want to  
go, method 1 or method 2 is quicker (see ex-  
amples).  
Press key until menu item ‘Preset  
counter’ is displayed:  
P R E S E T  
C O U N T E R  
d n - - -  
Example 1:  
= Preset counter activated  
P R E S E T  
C O U N T E R  
d n 0 5 6  
Actual preset counter reading = 50  
Former reference value = 300  
New reference value of 65 to be selected.  
= Preset counter deactivated  
Method 1 is most convenient:  
Preset counter function:  
Increase from 50 to 65 by pressing  
.
The preset counter allows the user to carry out  
a predefined number of sections in the ‘Continu-  
ous stroke’ sectioning mode.  
Example 2:  
Once the preselected number of sections has  
been carried out (= section counter reading ‘0’),  
the specimen head will stop automatically in the  
lower limit position.  
Actual preset counter reading = 50  
Former reference value = 300  
New reference value of 320 to be selected.  
Method 2 is most convenient:  
The reference value (i.e. the value from which  
the preset counter counts downward) can be  
selected in the corresponding menu item.  
Press  
reload old reference value (300):  
• Increase from 300 to 320 by pressing  
and  
simultaneously to  
.
In order to be able to set a reference value, the  
preset counter function must be active.  
For further information on the preset counter,  
please refer to chapter 5.3.6, Item 4 ‘Preset  
counter’.  
If the preset counter is deactivated (see diagrams  
above), press  
to activate.  
Instruction Manual V1.2 - 11/2006  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
The automatic defrost cycle takes place once  
every 24 hours at the selected defrost time.  
4. Quick-freezing at maximum specimen  
temperature* (instruments with specimen  
cooling only)  
7. Duration of defrost cycle:  
Press key until menu item ‘Fast object  
freezing’ is displayed:  
Press key until menu item ‘Defrost period’  
is displayed:  
F A S T  
O B J  
O F F  
F R E E Z I N  
D E F R O S T  
P E R I O D  
C H A M B E R :  
0 8  
M I N  
or  
F A S T  
O B J  
O N  
F R E E Z I N  
Use defrost keys (  
or  
) to adjust de-  
frost time. (Only if necessary - see below).  
Press lock/unlock key to activate or de-  
activate (ON/OFF).  
Important information for setting the  
defrost parameters  
Starting time of defrost cycle:  
5. Real time  
We recommend to program the  
automatic defrost cycle for a time out-  
side regular working hours as a  
period of 4 hours should pass after the  
defrost cycle before work with the  
instrument is resumed (see also  
chapter 5.7.1 ‘Automatic chamber  
defrost cycle’).  
Press key until menu item ‘Time’ is dis-  
played:  
T I M E  
1 0 : 2 3  
Use arrow keys (  
or  
) to set real time.  
Setting minutes as well as hours in  
1-minute steps. Holding arrow button in-  
creases scrolling speed!  
Defrost cycle duration:  
The duration of the chamber defrost  
cycle is set to 8 minutes in factory.  
This setting should only be  
changed, if the evaporator is not  
completely de-iced during the  
8-minute cycle.  
In such cases, increase the defrost  
cycle duration gradually until the the  
evaporator is ice-free after defrosting.  
The selected defrost cycle duration is  
valid for both automatic and  
manual chamber defrost cycle.  
6. Starting time of automatic defrost cycle:  
Press key until menu item ‘Defrost time’ is  
displayed:  
D E F R  
T I M E  
4 : 3 0  
(Time on display = start time of defrost  
cycle)  
Use arrow keys (  
sired defrost start time.  
or  
) to select the de-  
*) with 90° prism - see also chapter 5.5.2.  
Leica CM3050 S – Cryostat  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.3.5 Status display  
Each individual parameter is displayed in a certain area of the display screen.  
Upper line:  
In the menu (see chapter 5.3.4): indication of selected menu item.  
Standard indication: Chamber temperature (instruments with specimen cooling: chamber and specimen temperature)  
Standard indication upper line:  
Instruments with chamber and specimen cooling  
Instruments with chamber cooling only  
C T - 3 0 ° C  
C T - 3 0 ° C  
H O M E  
O T - 3 5 ° C  
ꢀꢀ  
H O M E  
-  
Upper line  
Lower line  
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
9
9
1 0 1 1 1 2 1 3 1 4 1 5 1 6  
1 0 1 1 1 2 1 3 1 4 1 5 1 6  
Digits 1 and 2  
indicate sectioning stroke position:  
Digits 3 - 6  
Digits 7 - 16  
indicate coarse feed position:  
indicate the following parameters:  
Sectioning stroke  
(outside sectioning window)  
• Rear limit position:  
• Front limit position:  
HOME  
STOP  
• Refrigeration status:  
Digit 15 = cryochamber  
Digit 16 = specimen  
Meaning of  
:
Retraction stroke  
Indication of ‘malfunction’ or  
‘handwheel locked’:  
-
Refrigeration OFF  
[ ]:  
Sectioning stroke  
refrigeration  
symbols:  
(inside sectioning window)  
• Microtome plug loose,  
implausible values etc.:  
Refrigeration ON  
Defrosting active  
????  
Sectioning  
stroke  
• Handwheel is locked:  
LOCKED  
• Section thickness  
(set value)  
Indication ‘LOCKED’ completely  
overlays digits 3 through 16,  
Sectioning window [  
]
i.e. while ‘LOCKED’ is displayed, no  
other parameters will be displayed  
in digits 3 - 16 of the lower line.  
• Section counter  
(actual value)  
• Indication of total section thickness  
(actual value)  
• Further indications: see chapter 6.1  
‘Error messages indicated in the  
display screen’  
ꢀꢀ  
Retraction stroke  
• Preset counter  
Digits 7 - 16:  
Scroll via arrow buttons to display  
individual parameters  
Instruction Manual V1.2 - 11/2006  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.3.6 Indication of refrigeration state, section thickness setting and counters  
These parameters can be displayed via the ar-  
row keys.  
4. Total section thickness counter  
C T - 3 0 ° C  
O T - 3 5 ° C  
μ 0 0 0 0 . 0  
When pressing  
played in the sequence listed on this page;  
when pressing , they are displayed in re-  
, the parameters are dis-  
Can also be set to ‘0’ at any time (via  
key).  
verse order. The display can be scrolled contin-  
uously, i.e. after the last parameter the display  
will automatically start again with the first.  
If the total section thickness indication is not set to  
‘0’ manually, the instrument counts up to maximum  
9999.5 μm at which point the counter will automati-  
cally switch back to ‘0’.  
This indication shows the total thickness of all  
sections carried out since the counter was last  
set to ‘0’. (All sections, in sectioning and trim-  
ming mode, are counted).  
1. Refrigeration state  
C T - 3 0 ° C  
H O M E  
O T - 3 5 ° C  
ꢀꢀ  
Meaning of the refrigeration symbols (asterisks):  
--> see chapter 5.3.5 ‘Status indication’.  
5. Preset counter  
C T - 3 0 ° C  
O T - 3 5 ° C  
d n 0 5 6  
2. Section thickness  
C T - 3 0 ° C  
O T - 3 5 ° C  
μ 0 1 0 . 0  
Setting preset counter to ‘0’:  
Press  
key.  
The sectioning thickness is set with the  
key (see chapter 5.4.7 ‘Setting the section  
thickness’).  
or  
can also be pressed while the continuous  
sectioning mode is in motion; in this case the  
preset counter is set to ‘0’ and sectioning stops  
upon reaching the lower limit position.  
After it has been set to ‘0’, the preset counter is  
inactive.  
3. Section counter  
C T - 3 0 ° C  
O T - 3 5 ° C  
n 0 0 0  
Reloading the former reference value:  
Once displayed (via arrow keys), the section  
counter can be set to ‘0’ at any time.  
Press  
and  
simultaneously.  
Reloading the former reference value is possible  
also while the continuous sectioning mode is in  
motion:  
Upon pressing both arrow keys, the instrument  
continues sectioning without interruption, start-  
ing to count down from the old reference value.  
To set to ‘0’ press  
.
If the section counter is not set to ‘0’ by the user  
at some point, sections are counted up to ‘999’  
at which point the counter automatically  
switches back to ‘0’.  
Leica CM 3050 S – Cryostat  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.4 Control panel 2  
Instruments without sectioning motor  
Instruments with sectioning motor  
In instruments without sec-  
tioning motor, control panel  
2 only consists of functional  
area B (see below)  
In instruments with section-  
ing motor, the control panel  
is equipped with three func-  
tional areas: A, B and C (see  
large diagram).  
A
Functional area A:  
• Sectioning modes  
• Sectioning window  
• Sectioning speed  
• Emergency stop function  
B
Functional area B:  
• Coarse feed  
• Trimming  
B
Functional area B:  
• Coarse feed  
• Trimming  
• Section thickness selection  
• Section thickness selection  
C
• Start / Stop sectioning  
Functional area C:  
Instruction Manual V1.2 - 11/2006  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Functional areas A and C - Instruments with sectioning motor only  
5.4.1 Selecting a sectioning mode /  
start and stop sectioning  
Start / stop motorized sectioning  
Sectioning mode 1 - Intermittent stroke:  
Three different sectioning modes can be select-  
ed for motorized sectioning:  
Simultaneously press and hold RUN/STOP and  
RUN/ENABLE or depress and hold footswitch.  
To stop the sectioning stroke, release the  
keys or the footswitch:  
Sectioning mode 1 - Intermittent stroke:  
The specimen head travels up and  
down as long as the RUN/STOP and  
RUN/ENABLE keys are pressed simulta-  
neously or as long as the footswitch is  
depressed.  
The specimen head stops immediately.  
Sectioning mode 2 - Continuous stroke:  
Simultaneously press and release RUN/STOP  
and RUN/ENABLE or depress and release  
footswitch.  
Sectioning mode 2 - Continuous stroke:  
The specimen head travels up and down  
until it is stopped by pressing RUN/STOP  
or RUN/ENABLE or by depressing the  
footswitch.  
To stop sectioning:  
Press and release RUN/STOP or RUN/EN-  
ABLE or depress and release the footswitch:  
The specimen head stops upon reaching the  
lower limit position.  
Sectioning mode 3 - Single stroke:  
The specimen head carries out one sec-  
tioning stroke and then stops in the low-  
er limit position.  
Sectioning mode 3 - Single stroke:  
Simultaneously press and release RUN/STOP  
and RUN/ENABLE or depress and release  
footswitch.  
Activating a sectioning mode  
The instrument carries out one complete  
section / retraction stroke and the specimen  
head stops upon reaching the lower limit po-  
sition.  
Press one of the three sectioning mode keys de-  
scribed above. While a sectioning mode is ac-  
tive, the LED in the corresponding key will be il-  
luminated.  
While sectioning is in motion, LED ‘RUN’ is illu-  
minated.  
While the sectioning motor is deactivated, LED  
‘STOP’ is illuminated.  
Both LEDs are located in functional area C.  
Leica CM 3050 S – Cryostat  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Functional areas A and C - Instruments with sectioning motor only  
The value has been stored when the LED in the  
sectioning window key lights up briefly.  
5.4.2 Setting the sectioning window  
The term ‘Sectioning window’ stands for a dif-  
ferent (slower) speed setting in that part of the  
sectioning stroke where the specimen is actual-  
ly cut, while the remaining sectioning stroke  
(above and below the specimen) as well as the  
retraction (return) stroke are carried out at a  
higher speed level (time-saving feature).  
If the LED flashes, the value has not yet been  
stored, because the user has tried to set a win-  
dow margin outside the possible window range  
(the area in which the sectioning window can  
be set is slightly smaller than the total section  
stroke).  
• In that case, repeat the step in question.  
.
If a sectioning window is set, speed inside the  
window is as set on the control panel. Outside  
the sectioning window the specimen head trav-  
els at a proportionately higher speed.  
The sectioning window cannot be  
deleted. If need be, it can be set to  
maximum size (i.e. almost the en-  
tire sectioning stroke will be car-  
ried out at one and the same  
speed).  
Always set the sectioning window  
while in the sectioning stroke (speci-  
men moves downward). Setting the  
window in retraction stroke (speci-  
men travels upward) is not possible.  
Every time after the instrument is  
disconnected from mains, the sec-  
tioning window is set to maximum  
size upon switching back on. This  
also applies when the instrument is  
switched on for the first time.  
Position lower edge of specimen approx.  
3 mm above the knife edge (rotate hand-  
wheel clockwise).  
Press sectioning window key.  
Position upper edge of specimen just below  
the knife edge (continue to rotate the hand-  
wheel clockwise).  
Press sectioning window key once  
more.  
Instruction Manual V1.2 - 11/2006  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Functional areas A and C - Instruments with sectioning motor only  
5.4.4 Emergency stop  
5.4.3 Selecting sectioning speed  
Activate by:  
Sectioning speed is selected via the sliding  
potentiometer. Speed can be changed while  
sectioning is in motion.  
either pressing the emergency stop  
button forcefully (must lock in  
position),  
or depressing the footswitch force-  
fully.  
Either way, sectioning is interrupted at once.  
Caution when working in continuous mode:  
If the slider is set to ‘0’, the sectioning motion  
stops, however, the actual sectioning mode is  
still active - just operating at ‘0’ speed.  
In both cases, the red LED in functional  
area A lights up, indicating that the emergency  
stop function has been activated. Simultaneously  
3
(= all 3 LEDs off).  
If the slider is accidentally moved, sectioning  
will be resumed! - Risk of injury!  
all  
sectioning modes are deactivated  
Therefore:  
To deactivate the emergency stop function,  
rotate the button in the direction of the ar-  
row until it unlocks and slips upward into its  
original position.  
Do not interrupt sectioning / trim-  
ming by setting the sliding potenti-  
ometer to zero speed!  
As soon as the emergency stop button has been  
unlocked, the red LED is extinguished.  
See also safety instruction in Chap-  
ter 2.3.6 ‘Motorized sectioning’.  
If the emergency stop function has been ac-  
tivated with the footswitch, unlocking is not  
necessary.  
Vmax key  
As soon as the footswitch is released, the red  
LED is extinguished.  
As long as the button is pressed, the  
sectioning motion is carried out at maxi-  
mum speed (210 mm/s).  
To continue work, select one of sectioning  
modes 1 - 3 and restart the sectioning  
motion.  
While Vmax is active, the sectioning window is  
inactive.  
See also chapter 2.2.3 ‘Safety - Emergency stop  
function’ for further information on the emer-  
gency stop.  
When the Vmax key is released, the sectioning  
window becomes active again.  
Leica CM 3050 S – Cryostat  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Functional area B - all models  
Coarse feed ‘fast forward’  
- Travelling speed 1,000 μm/s  
5.4.5 Coarse feed keys  
The four coarse feed keys in functional area B  
function as follows:  
To activate the function, press  
and hold.  
The specimen head travels forward (= to-  
wards the knife) as long as the key is held.  
Coarse feed ‘fast backward’  
- Travelling speed 1,000 μm/s  
Coarse feed ‘slow forward’  
- Travelling speed 500 μm/s  
Press  
key to activate.  
Specimen head travels backward  
(= away from knife):  
To activate the function, press  
and hold.  
a. until it reaches the rear limit position,  
The specimen head travels forward  
(= towards the knife) as long as the key is  
held.  
b. until it is stopped by pressing one of the  
coarse feed keys (to stop the function, any  
of the four coarse feed keys can be  
pressed).  
‘STOP’-LED (located in key  
‘Coarse feed fast forward’)  
Coarse feed ‘slow backward’  
- Travelling speed 500 μm/s  
As long as the forward coarse feed (both fast  
and slow) is in motion, the ‘STOP’ LED is illumi-  
nated.  
To start, press  
and hold.  
Specimen head travels backward (= away  
from knife) as long as the key is held.  
Upon reaching the front limit position, the  
‘STOP’ LED starts flashing and the display in  
control panel 1 reads ‘STOP’.  
LED ‘HOME’ (located in key  
‘Coarse feed fast backward’)  
As long as the backward coarse feed (both fast  
and slow) is in motion, the ‘HOME’ LED is  
illuminated.  
Upon reaching the rear limit position, the  
‘HOME’ LED starts flashing and the display in  
control panel 1 reads ‘HOME’.  
Instruction Manual V1.2 - 11/2006  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Functional area B - all models  
5.4.7 Setting the section thickness  
5.4.6 Trimming and sectioning functions  
Switching between trimming  
and sectioning  
Activate sectioning mode (  
off).  
C T . . . . ° C  
O T . .  
μ 0 1 0 . 0  
. . ° C  
Press this key to alternate between the  
two functions.  
Press  
or  
to select section thickness.  
illuminated:  
Trimming mode is active  
Pressing one of these keys once makes the  
display in control panel 1 switch to section  
thickness indication (thickness value not yet  
off:  
Sectioning mode is active.  
being altered). Pressing  
or  
further  
starts to change the actual section thick-  
ness setting.  
Selecting trimming thickness  
Activate trimming mode  
Select trimming thickness (= sectioning  
thickness at which trimming is carried out)  
Selectable section thickness settings:  
0
2
-
-
-
-
2 μm in 0.5-μm steps,  
by pressing  
or  
.
10 μm in 1-μm  
20 μm in 2-μm  
60 μm in 5-μm  
steps,  
steps,  
steps,  
steps,  
steps.  
10  
20  
The selected thickness is displayed in the cor-  
responding LED  
:
60 - 100 μm in 10-μm  
100 - 300 μm in 50-μm  
Motorized trimming  
Select ‘Continuous stroke’ in functional area  
A.  
Activate trimming by pressing RUN/STOP  
and RUN/ENABLE (located in functional  
area C) simultaneously or by depressing the  
footswitch.  
To stop trimming press RUN/STOP or RUN/  
ENABLE or depress the footswitch again.  
Trimming thickness can be changed while trim-  
ming is in motion by pressing  
or  
.
Leica CM 3050 S – Cryostat  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.5 Daily operation of the instrument  
5.5.1 Freezing specimens onto specimen  
discs with quick-freeze shelf and  
stationary heat extractor  
Select chamber temperature and, in instru-  
ments with specimen cooling, specimen  
temperature according to the type of speci-  
men to be sectioned (see also chapter 6.3  
‘Reference chart for temperature settings’).  
Do not continue with step 5.5.1 before the  
actual chamber / (specimen) temperature  
corresponds to the set value. This may take  
several hours (e.g. when operating the in-  
strument for the first time or when the instru-  
ment had been switched off).  
Size the specimen.  
Apply sufficient amount of embedding medi-  
um onto ambient temperature specimen  
disc.  
Inside the cryochamber, place the specimen  
onto the disc and orient.  
Place specimen disc with specimen into one  
of the 10 openings in the quick-freeze shelf  
(see below) and freeze.  
Prior to operating the instrument ,  
please read and observe the fol-  
lowing safety instructions:  
‘Integrated safety devices’  
(Chapter 2.2) .  
‘Handling microtome knives  
/ blades’  
(Chapter 2.3.4)  
2
1
‘Knife guard /  
Handwheel lock’  
(Chapter 2.3.5)  
3
’Motorized sectioning’  
(Chapter 2.3.6)  
The freezing process can be sped up further by  
using the stationary heat extractor (1). Next to  
the low temperature stabilizer (2) (‘Parking  
station’), 3 specimen discs (3) can be placed  
within its pivoting range.  
‘Defrosting  
/ handling frozen tissue’  
(Chapter 2.3.7)  
‘Frozen instrument parts  
and accessories’  
(Chapter 2.3.8)  
Place the heat extractor (1) directly onto the  
specimen surface. After approx. 30 seconds  
the specimen is completely frozen.  
‘Infectious / radioactive material’  
(Chapter 2.3.9)  
Instruction Manual V1.2 - 11/2006  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
The specimen head is now automatically cooled  
down to the lowest temperature (-50 °C).  
5.5.2 Freezing specimens onto specimen  
discs with 90 ° prism  
(Instruments w/specimen cooling only)  
Once the specimen is frozen, remove the  
disc from the prism and store it in the quick-  
freeze shelf until inserting it into the speci-  
men head.  
In instruments with specimen cooling, the 90 °  
prism, which is inserted into the specimen head,  
is an additional option to freeze specimens at  
maximum temperature .  
Press  
Object Freezing’ and deactivate the function  
by pressing once again (see chapter  
5.3.4 ‘Menu functions’ item 4).  
again to activate menu item ‘Fast  
4
The specimen head temperature now automati-  
cally returns to the previously selected set val-  
ue.  
Loosen screw (3) at specimen head and re-  
move prism (1).  
Installing the 90 ° prism  
Store prism inside the cryochamber.  
Loosen screw (3) at specimen head.  
Insert pin (1) of prism completely into recep-  
tacle (2) in the specimen head.  
5.5.3 Optional accessories for freezing  
specimens onto specimen discs  
and for removing frozen specimens  
Important: the entire rear surface of the  
prism must fit tightly against the specimen  
head surface!  
Mobile heat extractor  
Retighten screw (3) at specimen head.  
In addition to the stationary heat extractor (part  
of standard delivery), a mobile heat extractor is  
available as an optional accessory - see Ap-  
pendix ‘Ordering information - accessories ):  
The prism now forms a horizontal surface for  
quick specimen freezing.  
Freezing specimens with the 90 ° prism  
Apply sufficient amount of embedding medi-  
um onto ambient temperature specimen disc,  
add specimen and orient.  
Insert specimen disc with specimen into the  
receptacle (4) of the prism.  
Press to activate menu item ‘Fast Object  
Freezing’ and activate function by pressing  
Store heat extractor in chamber to precool.  
Place onto specimen in quick-freeze shelf.  
Remove as soon as specimen is frozen.  
(see chapter 5.3.4 ‘Menu functions’ item 4).  
Leica CM 3050 S – Cryostat  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.5.4 Inserting specimen discs into the  
specimen head  
Thermoblock  
The thermoblock makes it easy to remove fro-  
zen specimens from the specimen discs.  
Lock handwheel in upper position.  
If a knife has already been inserted  
into the knife holder, cover knife  
edge with knife guard!  
Do not store the thermoblock inside  
the cryochamber - store at room  
temperature and only place into the  
chamber for removing specimens  
from discs.  
Place cap (9) on one of the two ends so that  
the correct size opening (7) for the disc to be  
inserted is exposed.  
Loosen clamping screw (3).  
Insert specimen disc in opening (2) (rear  
surface of specimen disc must fit tightly  
against surface of specimen head).  
Insert pin (6) of specimen disc (5) into the  
opening (7).  
After about 20 seconds remove the frozen  
specimen from the specimen disc (use  
tweezers).  
Tighten clamping screw (3).  
If cap (9) fits too loosely, readjust with  
screw (10). Do not tighten screw too much!  
Once the specimen has been detached from  
the disc, remove the thermoblock from the  
cryochamber.  
5.5.5 Specimen orientation  
Open clamping lever (4) --> (4a).  
Orient specimen via orienting screws (1)  
and (5).  
Retighten clamping lever (4).  
Instruction Manual V1.2 - 11/2006  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.5.6 Trimming / sectioning  
Inserting the knife  
Use motorized coarse feed to bring speci-  
men close to knife (see chapter 5.4.5 ‘Coarse  
feed functions’).  
Move anti-roll plate away from knife.  
Insert the knife / disposable blade into the  
knife holder.  
-->Next steps: see chapter 5.4.6 ‘Trimming’.  
Select appropriate clearance angle at knife  
holder (settings between 4° and 6° are suit-  
able for the majority of applications).  
Sectioning  
Adjust knife / knife holder relative to speci-  
men (see ‘Knife holder’ instruction manual).  
For details on how to insert and ad-  
just the knives, please refer to the  
separate instruction manual(s) for  
your knife holder(s).  
See also chapter 2.2 ‘Integrated  
safety devices’ incl. subchapter  
2.2.4 ‘Knife guard’.  
Place anti-roll guide onto knife.  
Orient specimen, if necessary (see chapter  
5.5.5 ‘Specimen orientation’).  
Select desired section thickness (see chap-  
ter 5.4.7 ‘Setting the section thickness’).  
Start sectioning at approx. 20 μm, reduce  
section thickness gradually until reaching  
the appropriate thickness for the type of  
specimen to be sectioned.  
Manual trimming  
Adjust knife / knife holder relative to speci-  
men (see ‘Knife holder’ instruction manual).  
Discard the first 2 to 3 sections cut at the de-  
sired section thickness.  
Unlock handwheel.  
Use motorized coarse feed to bring speci-  
men close to knife (see chapter 5.4.5 ‘Coarse  
feed functions’).  
When sectioning manually, rotate the hand-  
wheel evenly and at uniform speed.  
Select desired section thickness (see chap-  
ter 5.4.7 ‘Setting the section thickness’).  
For all details on motorized sectioning, please  
refer to chapters 5.4.1 to 5.4.4:  
Move anti-roll plate away from knife.  
‘Selecting a sectioning mode / Start and stop  
sectioning’, ‘Setting the sectioning window’,  
‘Selecting sectioning speed’, ‘Emergency stop’.  
Rotate handwheel to trim specimen down to  
desired sectioning plane.  
Motorized trimming  
For advice on appropriate Sectioning tempera-  
tures for different types of tissue, please refer to  
chapter 6.3 ‘Reference chart for temperature  
settings’.  
Adjust knife / knife holder relative to speci-  
men (see ‘Knife holder’ instruction manual).  
Unlock handwheel.  
Center handwheel grip (See chapter 2.2.2).  
Leica CM 3050 S – Cryostat  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.6 Finishing work  
5.6.2 Switching off the instrument for a  
longer period of time  
5.6.1 Daily routine steps  
If you do not plan to use the instrument for an  
extended period of time, it can be switched off.  
Remove all specimens from the cry-  
ochamber and store safely! See  
also corresponding safety instruc-  
tions in chapter 2.3.7 ‘Defrosting /  
handling frozen tissue’!  
However, since this involves a lot of work (see  
below), it should be considered carefully  
whether switching off the instrument is really  
the better option.  
Lock handwheel.  
If the instrument is switched off, the microtome  
must be removed. Then the microtome, cryo-  
chamber and all accessories must be disinfect-  
ed and dried!  
Remove knife from knife holder, disinfect,  
dry (see chapter 7 ‘Disinfection, cleaning  
and maintenance’) and store in the knife  
case.  
Use a cold brush to sweep frozen sectioning  
waste into the section waste tray.  
For that purpose, follow the instructions in  
chapters 7.1 through 7.5.3.  
Empty section waste tray (dispose of sec-  
tioning waste according to the applicable  
regulations on disposal of special category  
waste!).  
If necessary, carry out a complete disinfection  
procedure (see chapter  
cleaning and maintenance’).  
7
‘Disinfection,  
Remove all specimens from the chamber  
and store safely.  
Close sliding window.  
Switch off cryochamber illumination.  
Use lock/unlock key to lock display in control  
panel 1 (Display background dark = locked):  
C T . . . . ° C  
O T .  
.
. . ° C  
1 2 : 0 0  
Do not switch off mains switch /  
automatic cutout (see chapter 5.2  
‘Switching on / fuses’)!  
Instruction Manual V1.2 - 11/2006  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
The specimen head refrigeration is automatical-  
ly switched back on when:  
5.7 Defrosting  
The Leica CM3050 S has 3 (instruments without  
specimen cooling: 2) different defrosting options.  
at least 4 hours have passed since the de-  
frost cycle was completed;  
and the actual chamber temperature devi-  
ates less than 5 K from set value.  
-
-
-
Automatic chamber defrost cycle  
Manual chamber defrost cycle  
After the specimen head refrigeration has been  
switched back on, the specimen head tempera-  
ture rises briefly to +10 °C and then decreases  
until it reaches the set value.  
Specimen head defrosting (manual) - instru-  
ments with specimen cooling only.  
Read the safety instructions in chap-  
ter 2.3.7 ‘Defrosting / frozen speci-  
mens’ !  
Switching specimen cooling back on immedi-  
ately after defrosting:  
As soon as the chamber defrost cycle is com-  
pleted, specimen cooling can be switched back  
on immediately:  
Make sure to remove sensitive speci-  
mens from the cryochamber before  
the defrost cycle starts!  
Press  
temperature’ is displayed:  
until menu item ‘Set specimen  
5.7.1 Automatic chamber defrost cycle  
S E T  
T E M P O T - 3 5 ° C  
The automatic defrost cycle is carried out once  
every 24 hours at the preselected time.  
(For details see chapter 5.3.4 ‘Menu functions’ -  
items 6 and 7).  
-  
Press  
.
S E T  
T E M P O T - 3 5 ° C  
ꢀꢀ  
During the defrost cycle, the corresponding as-  
terisk is displayed inversely:  
Specimen cooling is switched back on.  
Despite the possibility of switching  
the specimen cooling system back  
on immediately after the automatic  
chamber defrost cycle, we recom-  
mend the automatic defrost cycle  
be set at a point in time when the  
entire procedure can be controlled  
automatically.  
C T - 3 0 ° C  
H O M E  
O T - 3 5 ° C  
ꢀ  
Once the defrost cycle is finished, the instrument  
automatically switches back to refrigeration.  
Instruments with specimen cooling:  
When defrosting starts, to avoid icing the speci-  
men head is automatically adjusted to a temper-  
ature between -10 °C and -5 °C, i.e. the set tem-  
perature value is deactivated temporarily.  
(See also information on menu  
items 6 and 7 in chapter 5.3.4 ‘Menu  
functions’).  
Leica CM 3050 S – Cryostat  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.7.2 Manual chamber defrost cycle  
5.7.3 Defrosting the specimen head  
(instruments with specimen cooling  
only)  
In addition to the automatic defrost cycle, a  
manual chamber defrost cycle can be activat-  
ed when necessary.  
To activate specimen head defrosting:  
Press  
until menu item ‘Set specimen  
Press  
until menu item ‘Set chamber  
temperature’ is displayed:  
temperature’ is displayed:  
S E T  
T E M P O T - 3 5 ° C  
ꢀꢀ  
S E T  
T E M P  
C T - 3 0 ° C  
-  
Press  
. A brief sound signals that the  
Press  
. A brief sound signals that the  
manual chamber defrost cycle has started:  
manual chamber defrost cycle has started:  
S E T  
T E M P O T - 3 5 ° C  
S E T  
T E M P  
C T - 3 0 ° C  
ꢀ  
-  
Specimen head temperature increases to be-  
tween +20 °C and +30 °C for a duration of 10  
minutes, after which it returns to the set value.  
The manual defrost cycle ends either when the  
selected defrost cycle duration is over or when  
a chamber temperature of -5 °C is reached.  
To abort the manual defrost cycle:  
To abort specimen head defrosting:  
Go back to menu item ‘Set chamber tem-  
perature’ and press again.  
Press  
until menu item ‘Set specimen  
temperature’ is displayed and press  
once more.  
Instruments with specimen cooling:  
When defrosting starts, to avoid icing the spec-  
imen head is automatically adjusted to a tem-  
perature between -10 °C and -5 °C, i.e. the set  
temperature value is deactivated temporarily.  
(same as with automatic defrost cycle - see  
chapter 5.7.1 ‘Automatic chamber defrost cy-  
cle’)  
10 seconds after the manual chamber defrost cy-  
cle has been completed, the specimen head re-  
frigeration is switched back on, cooling down to  
the set value.  
Instruction Manual V1.2 - 11/2006  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Troubleshooting, applications tips  
6.1 Displayed error messages  
4. Service  
The autodiagnosis function of the CM3050 S  
displays error messages in case of certain mal-  
functions.  
C T - 3 0 ° C  
O T - 3 5 ° C  
S E R V I C E  
This message is displayed after a  
certain number of operating hours  
to remind the user to have a preven-  
tive maintenance done by Leica  
Technical Service.  
This message is displayed only, if  
the instrument is not regularly ser-  
viced by Leica Technical Service.  
1. Power failure  
This error message is displayed after the instru-  
ment had been disconnected from mains, either  
due to power failure or because the mains  
switch had been turned off.  
C T - 3 0 ° C  
P O W E R  
O T - 3 5 ° C  
F A I L U R E  
We recommend  
contract once the warranty period  
has expired!  
a
maintenance  
To acknowledge the message, press  
about 5 seconds.  
for  
2. Coarse feed  
Contact your local Leica sales  
office for details!  
Coarse feed malfunction has occurred.  
C T - 3 0 ° C  
? ? ? ?  
O T - 3 5 ° C  
If one of the error messages listed  
under 2, 3 and 4 is displayed, please  
contact Leica Technical Service.  
This message is also displayed, if the coarse  
feed plug has not been reconnected properly  
after reinstallation of the microtome.  
6.1.1 Other error indications  
3. Temperature indications  
(red ) in control panel 2 is illuminated:  
C T - # # ° C  
O T - 3 5 ° C  
Emergency stop button accidentally activat-  
ed.  
--> Unlock.  
or  
C T - 3 0 ° C  
O T - # # ° C  
Footswitch not properly connected.  
--> Check connection - connect properly,  
if applicable.  
If the # symbol is displayed instead of chamber  
or specimen temperature indication, the corre-  
sponding temperature sensor is broken or a  
short circuit has occurred.  
Leica CM 3050 S – Cryostat  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Troubleshooting, applications tips  
6.2 Potential problems - causes and remedies  
Problem  
Causes  
Remedies  
Frost on chamber walls and  
microtome  
- Cryostat is exposed to draft  
(open windows, doors, air-  
conditioning).  
- Eliminate draft sources or  
change place of installation of  
the cryostat.  
- Sliding window has been open  
for a long period of time at a  
very low chamber temperature.  
- Frost build-up caused by  
breathing into the cryocham-  
ber.  
- Close sliding window.  
- Consider wearing a mask.  
Sections smear  
- Specimen not cold enough.  
- Knife and/or anti-roll plate not  
yet cold enough - sections  
melt.  
- Select lower temperature.  
- Wait until knife and/or anti-roll  
plate have reached chamber  
temperature.  
Sections splinter  
Sections not properly flattened  
- Specimen too cold.  
- Select higher temperature.  
- Eliminate cause.  
- Static electricity / draft.  
- Specimen not cold enough.  
- Large surface specimen.  
- Select lower temperature.  
- Trim specimen parallel; in-  
crease section thickness.  
- Readjust anti-roll plate.  
- Align correctly.  
- Anti-roll plate poorly adjusted.  
- Anti-roll plate poorly aligned  
relative to knife edge.  
- Wrong clearance angle  
selected.  
- Knife blunt or damaged.  
- Set correct clearance angle.  
- Use different part of knife or  
replace knife.  
Sections not properly flattened  
despite correct temperature  
and correctly aligned anti-roll  
plate  
- Knife and/or anti-roll plate dirty.  
- Edge of anti-roll plate dam-  
aged.  
- Clean with dry cloth or brush.  
- Replace anti-roll plate.  
- Blunt knife.  
- Use different part of knife or  
replace knife.  
Instruction Manual V1.1 02/2000  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Troubleshooting, applications tips  
Problem  
Causes  
Remedies  
Sections curl on anti-roll plate.  
- Anti-roll plate does not pro-  
trude far enough beyond the  
knife edge.  
- Readjust correctly.  
Scraping noise during section-  
ing stroke and specimen return  
stroke.  
- Anti-roll plate protrudes too far  
beyond the knife edge and is  
scraping against the specimen.  
- Readjust correctly.  
Ondulated sections.  
- Knife damaged.  
- Use different part of the knife  
or replace knife.  
- Replace anti-roll plate.  
- Edge of anti-roll plate damaged.  
Chatter marks form during  
sectioning.  
- Specimen insufficiently frozen  
onto specimen disc.  
- Refreeze specimen onto disc.  
- Check disc clamping.  
- Specimen disc not clamped  
tightly enough.  
- Specimen holder ball joint not  
clamped.  
- Check ball joint clamping.  
- Check knife clamping.  
- Knife not clamped tightly  
enough.  
- Selected section thickness too  
thick - specimen has come off  
the specimen disc.  
- Refreeze specimen onto disc.  
- Specimen very hard and inho-  
mogeneous.  
- Increase section thickness; if  
possible, reduce specimen  
surface.  
- Blunt knife.  
- Use different part of knife or  
replace knife.  
- Knife profile inappropriate for  
the type of specimen to be  
sectioned.  
- Use knife with a different pro-  
file.  
- Wrong clearance angle.  
- Readjust clearance angle set-  
ting.  
Condensation forms on anti-roll  
plate and knife during cleaning.  
- Brush, forceps, cloth or other  
cleaning item too warm.  
- Accessories and cleaning  
material to be used pre-cooled  
only. - Store on storage shelves  
inside the cryochamber.  
Leica CM3050 S – Cryostat  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Troubleshooting, applications tips  
Causes  
Remedies  
Problem  
- Anti-roll plate protrudes too far  
beyond the knife edge. Adjust-  
ment was done in the direction  
of the knife.  
- Replace anti-roll plate; in  
future, lift away from the knife  
when aligning.  
- Handle anti-roll plate more  
carefully.  
Anti-roll plate damaged after  
adjustment.  
- Wrong temperature for the  
tissue to be sectioned.  
- Select correct temperature.  
- Wait until correct temperature  
has been reached.  
Thick/thin sections.  
- Knife profile inappropriate for  
the specimen to be sectioned.  
- Use knife with different profile  
(c or d) or possible switch to  
disposable blade system.  
- Remove ice.  
- Ice build-up on the knife back.  
- Handwheel rotation not uni-  
form or inappropriate rotation  
speed.  
- Adapt speed.  
- Knife not clamped tightly  
enough.  
- Specimen disc not clamped  
correctly.  
- Check knife clamping.  
- Check disc clamping.  
- Embedding medium poured  
onto cold specimen disc; spec-  
imen comes off the disc once  
frozen.  
- Place embedding medium on  
luke-warm specimen disc,  
inside the cryochamber place  
the specimen onto the disc and  
freeze in quick-freeze shelf.  
- Use different part of knife or  
replace knife.  
- Blunt knife.  
- Incorrect clearance angle.  
- Specimen dried out.  
- Set correct clearance angle.  
- Prepare new specimen.  
- Anti-roll plate too warm or  
incorrectly adjusted.  
- Static electricity.  
- Fat on corner or edge of anti-  
roll plate.  
- Cool or readjust anti-roll plate.  
Tissue sticks to the anti-roll  
plate.  
- Eliminate cause.  
- Remove fat (alcohol).  
- Knife rusty.  
- Remove rust.  
- Static electricity or draft.  
- Anti-roll plate too warm.  
- Eliminate cause.  
- Cool anti-roll plate.  
Flattened sections curl up as  
soon as anti-roll plate is lifted  
from the knife.  
Instruction Manual V1.1 02/2000  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Troubleshooting, applications tips  
Problem  
Causes  
Remedies  
Sections tear.  
- Temperature too low for the  
tissue to be sectioned.  
- Select higher temperature and  
wait for the new temperature  
to be reached.  
- Knife blunt, dirty, dusty, frosted  
or rusty.  
- Eliminate cause (--> see details  
on individual problems).  
- Exchange anti-roll plate.  
- Upper edge of anti-roll plate  
damaged.  
- Hard particles in the tissue.  
- If application permits, section  
at lower plane.  
- Clean.  
- Rear face of knife dirty.  
Inconsistent or insufficient  
specimen feed.  
- Microtome was not completely  
dry when refrigeration was  
- Remove microtome from cham-  
ber and dry - possibly with hair-  
dryer (hairdryer only to be used  
outside the cryochamber!) or at  
ambient temperature.  
switched on; ice build-up inside  
the micrometer mechanism.  
- Microtome defective.  
- Call Technical Service.  
Specimen disc cannot be  
removed.  
- Specimen disc stuck to quick-  
freeze shelf or specimen head  
due to moisture on underside  
of specimen disc.  
- Apply strong alcohol to contact  
point or heat the specimen  
head.  
Cryostat inoperational.  
- Mains plug not properly con-  
nected.  
- Check mains plug connection.  
- Automatic cut-out has been  
triggered.  
- Reconnect automatic cut-out  
(switch instrument back on).  
Sectioning motor inoperational.  
- Footswitch not connected.  
- Automatic cut-out of sectioning  
- Connect footswitch.  
- Reconnect automatic cut-out  
(push inwards until it locks in  
place).  
motor has been  
gered.  
- Motor defective.  
trig-  
- Call Technical Service.  
No or insufficient refrigerating  
performance.  
- Stopper not placed properly  
into drain in bottom of cryo-  
chamber.  
- Close drain with stopper.  
- Compressor defective.  
- Leak in cooling system.  
- Call Technical Service.  
- Call Technical Service.  
Leica CM3050 S – Cryostat  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Troubleshooting, applications tips  
6.3 Reference chart for temperature settings (in minus °C)  
Tissue  
10 - 15  
15 - 25  
25 - 50  
Adrenal  
Bone marrow  
Brain  
Bladder  
Breast - fatty  
Breast - little fat  
Cartilage  
Cervical  
Fatty  
Heart and vascular  
Intestinal  
Kidney  
Laryngeal  
Lip  
Liver  
Lung  
Lymphoid  
Muscular  
Nose  
Pancreatic  
Prostate  
Ovarian  
Rectal  
Skin with fat  
Skin without fat  
Spleenal or bloody tissue  
Testicular  
Thyroid  
Tongue  
Uterus curettage  
Please note: The recommended temperature settings are based on long-term experience. However, they can only be  
guidelines for general orientation, as each individual specimen may require particular adjustments.  
Instruction Manual V1.1 02/2000  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Disinfection, cleaning and maintenance  
7.2 Spray disinfection with Leica Cryofect  
7.1 Safety instructions  
on disinfection and cleaning  
For practical spray disinfection, we recommend  
Leica Cryofect which is fully effective at tem-  
peratures in the negative range down to -20 °C.  
When disinfecting the instrument,  
make sure to observe the following  
safety instructions:  
Set chamber temperature -20 °C or warmer.  
Remove specimens from the chamber and  
store safely.  
‘Disinfection and Cleaning’  
(Chapter 2.3.10)  
Carefully remove knife / blade from knife  
holder.  
‘Handling microtome knives /  
blades’  
(Chapter 2.3.4)  
Remove section waste tray from cryocham-  
ber and empty.  
‘Defrosting / handling frozen  
tissue’  
(Chapter 2.3.7)  
Remove remaining section waste from cryo-  
chamber.  
Working accessories (brushes, specimen  
holders etc.) can be disinfected inside the  
cryochamber.  
‘Frozen instrument parts and  
accessories’  
(Chapter 2.3.8)  
Once the selected chamber temperature has  
been reached, spray Leica Cryofect on the  
surfaces and accessories to be disinfected.  
‘Infectious / radioactive material’  
(Chapter 2.3.9)  
Allow to soak for 15 minutes.  
Use paper towels or another type of  
disposable cleaning cloths to wipe up  
disinfectant and remaining section waste.  
Dispose of paper towels together with sec-  
tion waste.  
If previously altered, set chamber tempera-  
ture back to its former value.  
Put specimens back into the cryochamber.  
Leica CM 3050 S – Cryostat  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7
Disinfection, cleaning and maintenance  
Allow cryochamber to dry thoroughly.  
Reinstall microtome (see chapter 7.5.3).  
Only when they are completely dry, reinsert  
the disinfected accessories / tools into the  
cryochamber.  
7.3 Disinfection with conventional  
disinfectants  
Switch the instrument off.  
Carefully remove knife / blade as well as spec-  
imens and accessories from the chamber.  
Remove section waste from the chamber  
and dispose of it.  
Reinsert the sliding window.  
Switch the instrument back on.  
Once a suitable chamber temperature has  
been reached, return the specimens to the  
cryochamber.  
Remove sliding window (see chapter 7.6 ‘Ex-  
changing the fluorescent light lamp’).  
Remove microtome (see chapter 7.5.1  
‘Removing the microtome’).  
Allow the microtome to thaw, disinfect and  
allow to dry thoroughly.  
7.4 Cleaning  
When necessary, the instrument surfaces  
can be cleaned with mild commercial house-  
hold cleaners.  
Drying the microtome:  
Use a hairdryer: dry and before  
reinstalling into the chamber  
allow the microtome to cool  
down to room temperature.  
Surfaces which are potentially contaminated  
(infectious germs) should always be cleaned  
with alcohol disinfectants!  
Allow to dry at room temperature  
(at least over night - longer,  
if necessary).  
Disinfect and dry disassembled parts  
(accessories and tools).  
Disinfect chamber walls with commercial  
alcohol disinfectant.  
Place an adequate container underneath the  
drain tube at the right side of the instrument.  
Pull stopper out of chamber bottom and  
drain off disinfecting solution into the con-  
tainer.  
Put stopper back into drain in cryochamber.  
Instruction Manual V1.2 - 11/2006  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Disinfection, cleaning and maintenance  
7.5 Removing / reinstalling the microtome  
7.5.2 Disassembling the specimen head  
(instruments w/specimen cooling only)  
7.5.1 Removing the microtome  
Use Allen key size 5 to remove screw (2)  
from the center of the specimen head (1).  
When removing the microtome  
make sure to observe the following  
safety instructions:  
‘Removing / reinstalling  
the microtome’  
(Chapter 2.3.11)  
‘Defrosting / handling frozen  
tissue’  
(Chapter 2.3.7)  
Do not twist the refrigerating tube (see  
chapter 2.3.11 ‘Removing / reinstalling the  
microtome’).  
Hold hook (4) (part of standard delivery) in  
one of your hands and with your other hand  
rest refrigerating tube (3) on hook.  
‘Infectious / radioactive material’  
(Chapter 2.3.9)  
‘Disinfection and cleaning’  
(Chapter 2.3.10)  
Insert hook (4) with refrigerating tube (3) into  
the opening above mark ().  
Prior to removing the microtome, carry out  
preparation steps as per chapter 7.3 (switch  
instrument off, remove accessories etc.)  
Leica CM 3050 S – Cryostat  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Disinfection, cleaning and maintenance  
Remove plug of temperature sensor (7).  
Remove plug of coarse feed motor:  
7.5.3 Reinstalling the microtome  
Pull bushing backward (arrow 1) and hold.  
Place your thumb between plug (6) and  
socket at the microtome (arrow 2).  
To remove plug (6) from microtome, pull it  
away from the microtome while at the same  
time carrying out a slight jogging movement  
When reinstalling the microtome  
make sure to observe the following  
safety instructions:  
‘Removing / reinstalling  
the microtome’  
(Chapter 2.3.11)  
(arrows 3).  
7
9
Display message ‘Dry Microtome’  
(Chapter 2.3.12)  
6
8
Lock handwheel in lower position.  
Place microtome in chamber.  
Put one drop of cryostat oil per surface on  
the contact surfaces of plastic coupling (11)  
and handwheel coupling (12).  
Place plastic coupling (11) on handwheel  
coupling (12).  
Push microtome slightly towards the right  
until axle (9) fits tightly into coupling (11).  
Unlock handwheel.  
Adjust the parts into each other by rotating  
the handwheel slightly backward and forth.  
Lock handwheel again in lower position.  
Tighten screws (8).  
Reconnect plugs of coarse feed motor (6)  
and temperature sensor (7).  
3
3
1
Loosen screws (8) with Allen key.  
Lift microtome a little and pull slightly to the  
left.  
This loosens the plastic coupling (11) which  
connects the two axles.  
Mount specimen head (in reverse order as  
described on previous page).  
Insert accessories (storage shelf, section  
waste tray etc.).  
Insert sliding window.  
Connect instrument to mains and switch  
back on.  
Remove the microtome from the chamber.  
Instruction Manual V1.2 - 11/2006  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Disinfection, cleaning and maintenance  
7.6 Exchanging the fluorescent light lamp  
Removal  
The lamp (3) is located behind the glare  
shield (4).  
Prior to exchanging the lamp,  
please read and observe the fol-  
lowing safety instructions:  
Carefully hold lamp with both of your hands.  
Slightly pull the lamp downward at the left  
side to remove it from fastening clip (5).  
Pull the lamp towards the left to remove it  
from its socket (6).  
‘Maintenance’  
(Chapter 2.3.13)  
Removing the sliding window  
Hold sliding window (1) at grip bar (2), lift  
slightly and pull out towards yourself.  
Installing the replacement lamp  
Hold the lamp in correct position (see draw-  
ing below) and push towards the right until it  
locks into the socket.  
Push the lamp slightly upwards until it locks  
into the clip.  
Put the sliding window back in place.  
Reconnect the instrument to mains and  
switch it back on.  
Leica CM 3050 S – Cryostat  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. Warranty and service  
Warranty  
Leica Microsystems Nussloch GmbH guarantees that the contractual prod-  
uct delivered has been subjected to a comprehensive quality control  
procedure based on the Leica in-house testing standards, and that the prod-  
uct is faultless and complies with all technical specifications and/or agreed  
characteristics warranted.  
The scope of the warranty is based on the content of the concluded agree-  
ment. The warranty terms of your Leica sales organization or the organiza-  
tion from which you have purchased the contractual product shall apply  
exclusively.  
Technical service information  
If you require technical service or replacement parts, please contact your  
Leica sales representative or dealer who sold the product.  
Please provide the following information:  
Model name and serial number of the instrument.  
Location of the instrument and name of the person to contact.  
Reason for the service call.  
Date of delivery.  
Decommissioning and disposal  
The instrument or parts of the instrument must be disposed of in compliance  
with the local laws.  
Instruction Manual V1.2 - 11/2006  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9. Appendix  
1. Electrohydraulic height adjustment  
1.3 Safety instructions  
An electrohydraulic lifting device for individual  
working height adjustment is available as an  
optional accessory for the CM3050 S cryostat  
series.  
Prior to operating the height adjustment device,  
make sure to read the present instruction manu-  
al. Pay special attention to the following safety  
instructions:  
1.1 Designated use  
Prior to connecting the height ad-  
justment device to mains, check to  
make sure that the nominal voltage  
and frequency indications on the  
device name plate correspond to  
the actual mains voltage in your  
lab!  
The electrohydraulic height adjustment device  
has been designed exclusively for adjustment  
of the instrument working level to the individual  
height of the user.  
It has been designed for use with the Leica cry-  
ostat series CM3050 S only and may under no  
circumstances be used to lift other instruments  
or persons.  
Do not move the height adjustment  
device with the Leica CM3050 S in-  
stalled on the frame. If the place of  
installation needs to be changed,  
first take the Leica CM3050 S down  
from the height adjustment device  
and then move both devices sepa-  
rately.  
Damage resulting from improper use of the  
product and/or faulty operation invalidates the  
warranty.  
Leica will not assume liability for any such dam-  
age.  
1.2 Standard delivery  
Never place your feet underneath  
the U-frame when lowering the  
height adjustment device! Risk of  
injury!  
Electrohydraulic height adjustment device for  
cryostat series CM3050 S, complete outfit, con-  
sisting of a U-shaped frame on four height-ad-  
justable feet and a two-key control unit.  
Only authorized Leica technical ser-  
vice engineers may carry out re-  
pairs on both the Leica CM3050 S  
and the height adjustment device.  
This applies especially to all work  
carried out on the hydraulic system  
of the height adjustment device.  
The height adjustment device is available in 2  
voltages:  
1. 230 – 240 V / 50 Hz – order no. 0443 26147  
2. 230 – 240 V / 50 Hz – order no. 0443 26147  
Leica CM 3050 S – Cryostat  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9.  
Appendix  
1.4 Installation  
Open the shipping box and remove all packing material including the Sty-  
rofoam pieces.  
Take the height adjustment device out of the box.  
At the place of installation, place the U-frame on the floor with the open  
side facing yourself. Adjust the U-frame parallel to the wall, maintaining a  
distance of approx. 10 cm between wall and U-frame.  
Push the (ab) button to lower the height adjustment device to the lowest  
possible position.  
In order to be able to move the instrument,  
use a fork wrench and unscrew the rollers  
on which the Leica CM3050 S moves by  
about 2 cm.  
Roll the cryostat onto the U-frame as far as it  
will go. The rear of the cryostat housing has  
to fit closely to the rear bar of the U-frame.  
Roll the cryostat onto the U-frame as far as it  
will go. The rear of the cryostat housing has  
to fit closely to the rear bar of the U-frame.  
When placing the instrument onto  
the U-frame, make sure the power  
cord does not get caught and thus  
crushed between U-frame and cry-  
ostat housing.  
Plug the height adjustment device into the  
mains socket. The unit is now ready for use.  
Instruction Manual V1.2 - 11/2006  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9. Appendix  
1.5 Control keys  
1.7 Troubleshooting  
1. Slight and uniform loss of height of the de-  
vice in the course of several weeks is nor-  
mal. Remedy:  
Correct the loss of height by pressing the  
button until the instrument is lifted back up to  
the desired working level.  
2. The cryostat is in an inclined position due to  
an uneven upward movement of the lifting  
pistons. Difference of level among the four  
legs > 5 mm. Remedy:  
Have the instrument travel once up and  
down over the entire height adjustment dis-  
tance: press  
until the instrument reach-  
until  
es the upper limit position and then  
the instrument reaches the lowest possible  
position. Then lift the instrument back to reg-  
ular working height.Should the problem re-  
occur several times, call Leica Technical  
Service.  
The control keys are located on top of the con-  
trol unit (see photograph).  
To lift the instrument, press  
.
3. The height adjustment device does not  
reach the maximum upper level or the instru-  
ment lowers itself by a relatively large dis-  
tance (> 5 cm) within a short time.  
To lower the instrument, press  
.
As long as a button is held, the instrument will  
go up or down until reaching the upper or lower  
limit.  
Remedy: Call Leica Technical Service.  
1.8 Technical Data  
Nominal voltage:  
Nomi nal frequency: 50 Hz  
230 - 240 V  
100 - 115 V  
60 Hz  
1.6 Maintenance and care  
Protect the lifting pistons in the four legs of  
the height adjustment device against dust  
and water.  
Type of actuation: Elektrohydraulik  
Total height of stroke: 230 mm  
Carrying power:  
Dimensions:  
Width:  
200 kp  
Approximately every six months lubricate the  
legs with a thin coat of resin-free oil.  
880 mm (without control unit)  
1030 mm (with control unit)  
555 mm  
Once every year have the height adjustment  
device checked by an authorized Leica ser-  
vice engineer.  
Depth:  
Weigth:  
approx. 30 kg  
Leica CM 3050 S – Cryostat  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9.  
Appendix  
2. Ordering information - accessories  
2.1 Knife holders  
Knife-holder CN for reusable microtome knives´w/anti-roll guide,  
adjustable knife guard and knife support bars for slim and wide knives ............................................. 0419 18649  
Knife support bar for knife holder CN, for slim knives .............................................................................. 0419 19426  
Knife support bar for knife holder CN, for wide knives ............................................................................ 0419 19427  
Knife-holder CE w/ glass anti-roll guide, for low profile disposable blades,  
with lateral displacement feature ................................................................................................................ 0419 30383  
Knife-holder CE w/ glass anti-roll guide, for high profile disposable blades  
with lateral displacement feature ................................................................................................................ 0419 30390  
Knife-holder CS for reusable knives, with one knife support  
(mounted optionally on the left or right), anti-roll guide and knife guard ............................................. 0419 19787  
Auxiliary clamping device for knife holder CS  
(to be used instead of knife support, for increased stability) ................................................................. 0419 20971  
Knife support bar for knife holder CS .......................................................................................................... 0419 25681  
Knife holder base for knife holders CN, CE and CS................................................................................... 0419 26140  
Knife rest with section collecting tray for knife holder CS ...................................................................... 0419 18996  
2.2 Specimen discs  
Specimen disc, 20 mm ø ................................................................................................................................ 0370 08636  
Specimen disc, 25 mm ø ................................................................................................................................ 0416 19275  
Specimen disc, 30 mm ø ................................................................................................................................ 0370 08587  
Specimen disc, 40 mm ø ................................................................................................................................ 0370 08637  
Specimen disc, 55 mm ø ................................................................................................................................ 0419 26491  
Specimen disc, 50 x 90 mm ............................................................................................................................ 0419 26750  
other sizes on request  
2.2 Accessories  
90° prism ........................................................................................................................................................... 0443 25949  
Mobile heat extractor..................................................................................................................................... 0443 26836  
Thermoblock .................................................................................................................................................... 0398 18542  
Glass anti-roll guide, complete assy., for knife holder CE ....................................................................... 0419 30402  
Glass anti-roll guide upgrade kit,  
for retrofitting knife holders CN, CS (for C-profile knives) ...................................................................... 0419 30403  
2.4 Consumables  
OCT embedding compound for freezing tissue specimens onto specimen discs,  
bottle of 125 ml ................................................................................................................................................. 0201 08926  
Cryostat oil No. 407, bottle of 100 ml ............................................................................................................ 0336 06099  
Instruction Manual V1.2 - 11/2006  
62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10. EC Declaration of Conformity  
EC Declaration of Conformity  
We herewith declare, in exclusive responsibility, that the  
Leica CM3050S – Cryostat  
was developed, designed and manufactured to conform with the  
Council Directive 73/23/EEC (Low Voltage),  
Council Directive 89/336/EEC, Appendix I (Electromagnetic Compatibility) and  
European council Directive 98/79/EG (IVDD).  
The following harmonized standards were applied:  
EN 61010-1: 2001  
Safety requirements for electrical equipment for measurement, control and laboratory use -  
Part 1: General requirements  
EN 61010-2-101: 2002  
Safety requirements for electrical equipment for measurement, control and laboratory use -  
Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment  
EN 61326-1: 1997 + A1: 1998 + A2: 2001 + A3: 2003  
Electrical equipment for measurement, control and laboratory use - EMC requirements -  
Part 1: General requirements  
EN 61000-3-2: 2000  
Electromagnetic compatibility (EMC)  
Part 3-2: Limits - Limits for harmonic current emissions  
EN 61000-3-3: 1995 + A1: 2001  
Part 3: Limits -  
Section 3: Limitation of voltage fluctuations and flicker in low-voltage  
supply systems for equipment with rated current £ 16 A  
Electromagnetic compatibility (EMC)  
EN 418: 1992  
Emergency stop equipment.  
Principles for design.  
EN 1037: 1995  
Safety of machinery.  
Prevention of unexpected start-up.  
In addition, the following in-house standards were applied:  
DIN EN ISO 9001: 2000.  
Leica Microsystems Nussloch GmbH  
Postfach 1120  
D-69222 Nussloch  
. . . . . . . . . . . . . . . . . . . . . . . . . . . .  
Dag Graupner  
Managing Director  
January 12, 2004  
Download from Www.Somanuals.com. All Manuals Search And Download.  
11. Decontamination Certificate (master)  
Dear Customer,  
any product which is be returned to Leica Microsystems or serviced on site, must be cleaned and  
decontaminated in the appropriate manner. This certificate, duly completed, must be placed in the unit  
prior to packaging or handed over to the service engineer. Packages will not be opend nor servicing  
commenced until the Company or service engineer have received a satisfactory certificate. Should  
returned goods be considered a hazard by the Company, they will be returned immediately to the  
customer at his/her expense.  
NB: Microtome knives must be in boxes.  
Description  
Name/Model  
KAT No.  
Fabr. No.  
Quantity  
Tick Box A if applicable. Otherwise please complete all parts of B, providing  
further information as requested or appropriate.  
Yes  
This equipment has not been in contact with unfixed biological samples.  
A
B
1
This equipment has been exposed internally or externally to hazardous materials as  
indicated below:  
Yes No  
Please provide further detail here:  
Blood, body fluids, pathological samples  
Other biohazards  
Chemicals/substances hazardous to health  
Other hazards  
2
This equipment has been cleaned and decontaminated:  
If yes, give details of the method:  
Yes No  
Please provide further detail here:  
If no*, please indicate why not:  
* Such equipment must not be returned without the written agreement of Leica Microsystems.  
3
The equipment has been prepared to ensure safe handling/transportation.  
Whenever possible, please use the original transportation case/box.  
Yes No  
Instruction Manual V1.2 - 11/2006  
64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
11. Decontamination Certificate (master)  
Important - to avoid refusal of shipment:  
Place one copy in the unit prior to packaging or hand it over to the service engineer. Customer as-  
sumes all responsibility for the immediate return shipment of articles sent to Leica without proper  
decontamination documentation.  
If you have any further question, please call your local Leica organisation.  
Leica Internal Use: If applicable, note corresponding Job and RAN-/RGA-Number:  
Job Sheet No.:  
____________________  
SU Return Goods Authorisation: ____________________  
BU Return Authorisation Number: ____________________  
Signature/Date  
Institute  
Department  
Adress  
Name  
Position  
eMail  
Phone  
Facsimile  
Leica Microsystems Nussloch GmbH  
Heidelberger Str. 17-19  
69226 Nussloch, Germany  
Phone:++49 (0) 6224 143 0  
Fax: ++49 (0) 6224 143 200  
Leica CM 3050 S – Cryostat  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Instruction Manual V1.2 - 11/2006  
66  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Leica Tablet Accessory TPS400 User Manual
Lenovo Laptop 2394F1U User Manual
LG Electronics Cell Phone 355 User Manual
LG Electronics Home Theater System HB906TA User Manual
Lincoln Electric Welder UWW 170 User Manual
Mackie Microphone MT 3100 User Manual
Mattel Doll Y2569 User Manual
McCulloch Snow Blower 96182000500 User Manual
Meco Charcoal Grill 2000 Brazier User Manual
Meyer Furnace 2900 User Manual