Konica Minolta Network Card LED 850 User Manual

LED-850  
TEC Test Adapter for High-Power LEDs  
Two Global Leaders.  
One Complete Solution.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Your total light and display sensing solution.  
Product Highlights  
Temperature range from +5°C to +85°C (TEC heating  
and cooling)  
Interchangeable sockets for all common high-power  
LEDs and small Level 2 printed circuit boards  
Accuracy to 0.1°C  
Maximum LED power rating of 10 W  
High-Power Also Means  
High-Temperature  
The temperature of the LED’s chip has an enormous influence  
on the light output emitted and on the radiation spectrum  
(LED color). This interdependency must be taken into account  
particularly in the case of high-power LEDs because the high  
operating temperatures lead to an undesirable deterioration in  
performance (i.e. reduction in their light output). To be able to  
integrate high-power LEDs successfully in final applications, it  
is therefore essential to understand the optical properties as a  
function of the temperature.  
Modular Design for Flexible Usage  
The modular system consists of the basic LED-850 unit with  
Peltier element and heat sink, and a socket (with centering cap)  
specifically matched to the respective model of high-power  
LED. A mechanical locking lever locks the centering cap securely  
in place, guaranteeing correct mounting of the LED every time.  
The centering cap also provides the electrical connection.  
Two types of centering caps are available:  
A sand-blasted, white anodized cap for measuring the  
luminous flux using integrating spheres (this centering cap  
belongs to the scope of delivery of the LED socket)  
A black anodized cap for measuring the luminous intensity  
using CIE compliant, LED25-xxx series averaged LED  
intensity adapters (cap optional)  
Measuring Thermal Behavior with  
the LED-850  
TEC test adapter LED-850 enables the influence of electrical  
and thermal operating conditions on the optical parameters  
of high-power LEDs to be assessed quickly and easily. A  
Peltier element with cooler fan is used to cool or heat the test  
LED within a range between +5 and +85°C. What’s more,  
the integral PT100 sensor and the Peltier element can be  
connected to a TEC controller that controls and displays the  
temperature. Robust test probes also ensure reliable electrical  
contacting of the LED.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Your total light and display sensing solution.  
The temperature of the LED is controlled by means of a gold-plated  
Rapid Change-Over of the  
LED Socket  
The LED socket can be changed over easily and quickly. The  
modular design principle makes it possible to test various types  
of LED using just a single base unit and interchangeable sockets.  
The range of LED sockets (with matching centering cap) covers  
all commonly available high-power LEDs from the world‘s leading  
LED manufacturers, as well as Level 2 printed circuit boards  
(e.g. Luxeon Star). Customer-specific adapters can also be  
supplied on request.  
copperzirconium rod that is pressed against the heat-conducting  
slug of the LED’s housing by springs. The high thermal conductivity  
of the copper-zirconium rod and the Peltier element – with maximum  
cooling/heating capacity in excess of 120 W – allow the temperature  
to be set quickly to the desired value within a range from +5 to  
+85°C. In addition, a large heat sink and a fan assist efficient  
dissipation of the heat loads. This results in temperature control  
times of just a few minutes. What’s more, a PT100 temperature  
sensor that guarantees setting accuracy down to 0.1°C is housed in  
the copper-zirconium rod beneath the contact surface for the LED.  
Setting time from ambient temperature (Tu=23°C and 35°C) to 15°C; LED power  
dissipation 5 W; Peltier power rating 50W; PID parameters: P=80, I=0.3, D=0  
Precise 4-Wire System for  
Electrical Contacting  
Electrical contacting of the LED is provided by a four-wire  
system, a set of test probes and contact springs in the  
centering cap. This enables not only the operating current to be  
applied but also the forward voltage to be measured precisely  
at the same time. TEC test adapter LED-850 is rated for LED  
power ratings of up to 10 W. The test probes can easily be  
changed by users themselves.  
Efficient Heat Transfer Using A  
Copper-Zirconium Rod  
Power Supply and TEC Controller  
Instrument Systems recommends the 5305 TECSource made  
by Arroyo Instruments for controlling the temperature. We have  
developed control software for this device that is included in the  
scope of delivery of the LED-850. This userfriendly software can  
be used to automate measurements. A Keithley 2400/2600 series  
sourcemeter is highly suitable as the current source. Instrument  
Systems has developed software for this device, too.  
Thermal simulation results: The LED is efficiently cooled and the heat at the Peltier  
element dissipated by means of the heat sink  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contact information  
Instrument Systems - Operating Globally, Acting Locally  
Instrument Systems is the premier company for light measurement. As a subsidiary company of Konica Minolta we have access to  
resources over the long term and to a global network – our mission is your solution, anytime and anywhere.  
World-class customer support is rewarded with customer loyalty - a top priority at Instrument Systems.  
Neumarkter Str. 83  
D-81673 München, Germany  
Phone: +49 (089) 45 49 43 - 0  
Fax: +49 (089) 45 49 43 - 11  
Konica Minolta Sensing Americas  
Konica Minolta Sensing, Americas represents Instrument Systems in North America and Canada with a high level of expertise in marketing  
optical measuring instruments. High-quality customer support is based on a track record of close cooperation and constant exchange of  
information and know-how.  
101 Williams Drive  
Ramsey, NJ 07446 USA  
Phone: 201-236-4300  
Toll Free: 888-473-2656  
Fax: 201-785-2480  
sensing.konicaminolta.us  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Your total light and display sensing solution.  
Ordering Information  
Order Number  
Description  
TEC test adapter LED-850  
Base unit with active TEC cooling and heating; including universal mount for LED adapters with 50-mm diameter; for LEDs and LED clusters  
with max. 10 W power dissipation; incl. PT100 temperature sensor and 4-wire connection; socket for specific LED package not included.  
LED-850-0  
Socket for LUMILEDS LUXEON EMITTER LEDs (Type: Batwing, Lambertian, High Dome); socket dimensions: 7.95 x 7.3 mm; incl. centering cap  
with 50 mm diam. for luminous flux measurment  
LED-851-1  
LED-851-1-3  
LED-851-2  
Centering cap with 50 mm diam. for luminous intensity measurement; for LUMILEDS LUXEON EMITTER socket LED-851-1  
Socket for LUMILEDS LUXEON K2 LEDs; socket dimensions: 7.5 x 7.5 mm; incl. centering cap with  
50 mm diam. for luminous flux measurment  
LED-851-2-3  
LED-851-5  
Centering cap with 50 mm diam. for luminous intensity measurement; for LUMILEDS LUXEON K2 socket LED-851-2  
Socket for LUMILEDS LUXEON STAR LEDs; centering cap with 50 mm diam. for luminous flux measurment  
Centering cap with 50 mm diam. for luminous intensity measurement; for LUMILEDS LUXEON STAR socket LED-851-5  
LED-851-5-3  
Socket for OSRAM OS GOLDEN DRAGON LEDs; socket dimensions: 7.1 x 6.1 mm; incl. centering cap with 50 mm diam.  
for luminous flux measurment  
LED-852-1  
LED-852-1-3  
LED-853-1  
Centering cap with 50 mm diam. for luminous intensity measurement; for OSRAM OS GOLDEN DRAGON socket LED-852-1  
Socket for CREE XLamps 7090 LEDs; socket dimensions: 7.2 x 9.2 mm; incl. centering cap with 50 mm diam. for luminous flux measurement  
Centering cap with 50 mm diam. for luminous intensity measurement; for CREE XLamps 7090 socket LED-853-1  
LED-853-1-3  
Socket for NICHIA JUPITER NCCx022 LEDs; socket dimensions: 7.4 x 8.9 mm; incl. centering capwith 50 mm diam.  
for luminous flux measurement  
LED-854-1  
LED-851-1-3  
Centering cap with 50 mm diam. for luminous intensity measurement; for NICHIA JUPITER NCCx022 socket LED-854-1  
Accessory  
W-110  
W-121  
W-210  
Keithley Sourcemeter 2400 (IEEE-488 interface for PC and interface cable are nor included)  
Keithley Sourcemeter 2601A; (IEEE-488 interface for PC and interface cabel are nor included)  
Arroyo Instruments 5305 TECSource, 60 W; USB and RS-232 interface  
Instrument Systems is continually working to develop and improve products. Any technical changes, errors or misprints  
do not form grounds for compensation. The company`s Terms of Delivery and Payment apply in all other respects.  
Technical Specifications  
Thermal specifications  
Temperature control accuracy  
Maximum temperature  
+/- 0.1°C  
85°C  
Minimum temperature  
5°C  
Temperature control time to obtain min. and max. temperatures  
Thermal resistance from LED slug to PT 100  
Electrical specifications  
8 min  
2 K/W  
Max. current of Peltier element  
Max. voltage of Peltier element  
Max. power dissipation of LED  
Operating conditions  
8.5 A  
14.4 V  
10 W  
Temperature range  
15°C – 35°C  
70% rF  
Max. humidity  
General specifications  
Outside dimensions (H, W, D)  
Weight  
135 x 140 x 160 mm  
2.5 kg  
In the interest of the continuous development and improvement of our products, we reserve  
all rights to make technical changes to the information contained in our printed material.  
Such changes, errata and printing mistakes do not constitute grounds for claims for  
damages. The LumiCam series of measurement cameras may only be used for luminance  
and chromaticity measurements of self-illuminating objects.  
SAFETY PRECAUTIONS  
For correct use and for your safety, be sure to read  
the instruction manual before using the instrument.  
Always connect the instrument to the specified  
power supply voltage. Improper connection may  
cause a fire or electric shock.  
Certificate No : LRQ 09600941A  
Registration Date : March 3, 1995  
Certificate No : JQA-E-80027  
Registration Date : March 12, 1997  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Indesit Washer Dryer WIDXXL 146 User Manual
Insignia Video Game Controller NS GPS4HA101 User Manual
Integra Stereo Receiver DTR 51 User Manual
Jacuzzi Hot Tub 6042 User Manual
Janome Sewing Machine 350E User Manual
JVC Speaker System UX QD90B User Manual
JVC Stereo System 0609YAMMDWBET User Manual
Kenwood Range CK 740 User Manual
Kenwood Stereo Amplifier KAC 821 User Manual
KidzMouse Portable Radio SMB 631 User Manual