Agilent Technologies Welding System 33120A User Manual

Service Guide  
Publication Number 33120-90017 (order as 33120-90104 manual set)  
Edition 6, March 2002  
© Copyright Agilent Technologies, Inc. 1994-2002  
For Safety information, Warranties, and Regulatory information,  
see the last page in this manual.  
Agilent 33120A  
15 MHz Function /  
Arbitrary Waveform Generator  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note: Unless otherwise indicated, this manual applies to all Serial Numbers.  
The Agilent Technologies 33120A is a high-performance 15 MHz  
synthesized function generator with built-in arbitrary waveform  
capability. Its combination of bench-top and system features makes this  
function generator a versatile solution for your testing requirements now  
and in the future.  
Convenient bench-top features  
œ 10 standard waveforms  
œ Built-in 12-bit 40 MSa/s arbitrary waveform capability  
œ Easy-to-use knob input  
œ Highly visible vacuum-fluorescent display  
œ Instrument state storage  
œ Portable, ruggedized case with non-skid feet  
Flexible system features  
œ Four downloadable 16,000-point arbitrary waveform memories  
œ GPIB (IEEE-488) interface and RS-232 interface are standard  
œ SCPI (Standard Commands for Programmable Instruments) compatibility  
œ Agilent IntuiLink Arb Waveform Generation Software for  
®
®
Microsoft Windows  
The procedures in this manual are intended for use by qualified,  
service-trained personnel only.  
Warning  
Agilent 33120A  
15 MHz Function /  
Arbitrary Waveform Generator  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The Front Panel at a Glance  
1
2
3
4
Function / Modulation keys  
Menu operation keys  
Waveform modify keys  
Single / Internal Trigger key  
(Burst and Sweep only)  
5
6
7
8
Recall / Store instrument state key  
Enter Number key  
Shift / Local key  
Enter Number “units” keys  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
Front-Panel Number Entry  
You can enter numbers from the front-panel using one of three methods.  
Use the knob and the arrow keys to modify the displayed number.  
Use the arrow keys to edit individual digits.  
Increments the flashing digit.  
Decrements the flashing digit.  
Moves the flashing digit to the right.  
Moves the flashing digit to the left.  
Use the “Enter Number” mode to enter a number with the appropriate units.  
Use “Enter” for those operations that do not  
require units to be specified (AM Level,  
Offset, % Duty, and Store/Recall State).  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
The Front-Panel Menu at a Glance  
The menu is organized in a top-down tree structure with three levels.  
A: MODulation MENU  
1: AM SHAPE 2: AM SOURCE  
3: FM SHAPE  
4: BURST CNT  
9: FSK RATE  
5: BURST RATE  
10: FSK SRC  
6: BURST PHAS  
7: BURST SRC 8: FSK FREQ  
B: SWP (Sweep) MENU  
1: START F  
2: STOP F  
3: SWP TIME  
4: SWP MODE  
*
C: EDIT MENU  
1: NEW ARB  
[ 2: POINTS ]  
[ 3: LINE EDIT ]  
[ 4: POINT EDIT ]  
[ 5: INVERT ]  
[ 6: SAVE AS ]  
7: DELETE  
The commands enclosed in square brackets ( [ ] ) are “hidden” until you make a selection from the  
NEW ARB command to initiate a new edit session.  
*
D: SYStem MENU  
1: OUT TERM  
2: POWER ON  
3: ERROR  
4: TEST  
5: COMMA  
6: REVISION  
E: Input / Output MENU  
1: HPIB ADDR  
2: INTERFACE  
3: BAUD RATE  
4: PARITY  
5: LANGUAGE  
*
F: CALibration MENU  
1: SECURED or [ 1: UNSECURED ]  
[ 2: CALIBRATE ]  
3: CAL COUNT  
4: MESSAGE  
The commands enclosed in square brackets ( [ ] ) are “hidden” unless the function generator  
is UNSECURED for calibration.  
*
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Display Annunciators  
Adrs  
Rmt  
Trig  
Function generator is addressed to listen or talk over a remote interface.  
Function generator is in remote mode (remote interface).  
Function generator is waiting for a single trigger or external trigger (Burst, Sweep).  
AM modulation is enabled.  
AM  
FM  
FM modulation is enabled.  
Ext  
Function generator is set for an external modulation source (AM, FSK, Burst).  
FSK (frequency-shift keying) modulation is enabled.  
Burst modulation is enabled.  
FSK  
Burst  
Swp  
ERROR  
Offset  
Shift  
Num  
Arb  
Sweep mode is enabled.  
Hardware or remote interface command errors are detected.  
The waveform is being output with an offset voltage.  
“Shift” key has been pressed. Press “Shift” again to turn off.  
“Enter Number” mode is enabled. Press “Shift-Cancel” to disable.  
Arbitrary waveform function is enabled.  
Sine waveform function is enabled.  
Square waveform function is enabled.  
Triangle waveform function is enabled.  
Ramp waveform function is enabled.  
To review the display annunciators, hold down the Shift key as you  
turn on the function generator.  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
The Rear Panel at a Glance  
1
2
3
Chassis ground  
Power-line fuse-holder assembly  
Power-line voltage setting  
5
External Trigger / FSK / Burst modulation  
input terminal  
GPIB (IEEE-488) interface connector  
RS-232 interface connector  
6
7
4
AM modulation input terminal  
Use the front-panel Input /Output Menu to:  
œ Select the GPIB or RS-232 interface (see chapter 4 in User’s Guide).  
œ Set the GPIB bus address (see chapter 4 in User’s Guide).  
œ Set the RS-232 baud rate and parity (see chapter 4 in User’s Guide).  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
In This Book  
Specifications Chapter 1 lists the function generator’s specifications  
and describes how to interpret these specifications.  
Quick Start Chapter 2 prepares the function generator for use and  
helps you get familiar with a few of its front-panel features.  
Front-Panel Menu Operation Chapter 3 introduces you to the front-panel  
menu and describes some of the function generator’s menu features.  
Calibration Procedures Chapter 4 provides calibration, verification,  
and adjustment procedures for the function generator.  
Theory of Operation Chapter 5 describes block and circuit level theory  
related to the operation the function generator.  
Service Chapter 6 provides guidelines for returning your function  
generator to Agilent for servicing, or for servicing it yourself.  
Replaceable Parts Chapter 7 contains a detailed parts lists of the  
function generator.  
Schematics Chapter 8 contains the function generator’s block diagram,  
schematics, disassembly drawings, and component locator drawings.  
For information on using the Phase-Lock Option for the 33120A, refer to  
the User’s and Service Guide included with the Option 001.  
If you have questions relating to the operation of the 33120A,  
call 1-800-452-4844 in the United States, or contact your nearest  
Agilent Technologies Sales Office.  
If you believe your 33120A has failed, refer to “Operating Checklist”,  
“Types of Service Available”, and “Repackaging for Shipment” at the  
beginning of chapter 6.  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Chapter 1 Specifications  
Chapter 2 Quick Start  
To prepare the function generator for use 21  
If the function generator does not turn on 22  
To adjust the carrying handle 24  
To set the output frequency 25  
To set the output amplitude 26  
To set a dc offset voltage 27  
To set the duty cycle 28  
To output a stored arbitrary waveform 29  
To output a dc voltage 30  
To store the instrument state 31  
To rack mount the function generator 33  
Chapter 3 Front-Panel Menu Operation  
Front-panel menu reference 37  
A front-panel menu tutorial 39  
To select the output termination 44  
To output a modulated waveform 45  
To unsecure the function generator for calibration 47  
Chapter 4 Calibration Procedures  
Agilent Calibration Services 51  
Calibration Interval 51  
Time Required for Calibration 51  
Automating Calibration Procedures 52  
Recommended Test Equipment 52  
Test Considerations 53  
Performance Verification Tests 54  
Frequency Verification 56  
Function Gain and Linearity Verification 56  
DC Function Offset Verification 57  
AC Amplitude Verification 57  
Amplitude Flatness Verification 60  
AM Modulation Depth Verification 61  
Optional Performance Verification Tests 62  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Chapter 4 Calibration Procedures (continued)  
Calibration Security Code 64  
Calibration Count 66  
Calibration Message 66  
General Calibration/Adjustment Procedure 67  
Aborting a Calibration in Progress 69  
Frequency and Burst Rate Adjustment 69  
Function Gain and Linearity Adjustment 70  
AC Amplitude Adjustment (High-Z) 70  
Modulation Adjustment 72  
AC Amplitude Adjustment (50W) 73  
DC Output Adjustment 76  
Duty Cycle Adjustment 77  
AC Amplitude Flatness Adjustment 77  
Output Amplifier Adjustment (Optional) 80  
Error Messages 81  
Chapter 5 Theory of Operation  
Block Diagram Overview 85  
Output Attenuator 86  
Output Amplifier 87  
AM Modulation 89  
Pre-attenuator 90  
Square Wave and Sync 90  
Filters 92  
Waveform DAC/Amplitude Leveling/Waveform RAM 93  
Direct Digital Synthesis (DDS ASIC) 95  
System DACs 96  
Floating Logic 97  
Earth-Referenced Logic 98  
Power Supplies 98  
Display and Keyboard 100  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents  
Chapter 6 Service  
Operating Checklist 103  
Types of Service Available 104  
Repackaging for Shipment 105  
Cleaning 105  
Electrostatic Discharge (ESD) Precautions 106  
Surface Mount Repair 106  
To Replace the Power-Line Fuse 107  
To Replace the Output Protection Fuse (F801) 107  
Troubleshooting Hints 108  
Self-Test Procedures 110  
Chapter 7 Replaceable Parts  
Replaceable Parts 113  
Chapter 8 Schematics  
33120A Block Diagram 129  
Mechanical Disassembly 130  
Floating Logic Schematic 131  
Digital Waveform Data Synthesis 132  
System DAC Schematic 133  
Waveform DAC Schematic 134  
Filters Schematic 135  
Sync, Square Wave, and Attenuator Schematic 136  
Output Amplifier Schematic 137  
Output Attenuator Schematic 138  
Earth Reference Logic Schematic 139  
Power Supplies Schematic 140  
Display and Keyboard Schematic 141  
33120-66521 Component Locator Diagram 142  
33120-66502 Component Locator Diagram 143  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1
1
Specifications  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 1 Specifications  
Agilent 33120A Function Generator  
SIGNAL CHARACTERISTICS  
WAVEFORMS  
Squarewave  
Rise/Fall Time:  
Overshoot:  
Asymmetry:  
Duty Cycle:  
Standard Waveforms:  
Sine, Square, Triangle,  
Ramp, Noise, DC volts,  
Sine(x)/x, Negative Ramp,  
Exponential Rise,  
< 20 ns  
< 4%  
1% + 5 ns  
20% to 80% (to 5 MHz)  
40% to 60% (to 15 MHz)  
Exponential Fall, Cardiac  
Arbitrary Waveforms:  
Waveform Length:  
Amplitude Resolution:  
Sample Rate:  
Triangle, Ramp, Arb  
Rise/Fall Time:  
Linearity:  
Settling Time:  
Jitter:  
8 to 16,000 points  
12 bits (including sign)  
40 MSa / sec  
40 ns (typical)  
< 0.1% of peak output  
< 250 ns to 0.5% of final value  
< 25 ns  
Non-Volatile Memory:  
Four 16,000-point waveforms  
FREQUENCY CHARACTERISTICS  
OUTPUT CHARACTERISTICS (1)  
Sine:  
100 Hz 15 MHz  
100 Hz 15 MHz  
100 Hz 100 kHz  
100 Hz 100 kHz  
10 MHz bandwidth  
(2)  
Amplitude (into 50 ):  
Accuracy (at 1 kHz):  
Flatness  
50 mVpp – 10 Vpp  
Square:  
1% of specified output  
(sine wave relative to 1 kHz)  
1% (0.1 dB)  
Triangle:  
Ramp:  
Noise (Gaussian):  
Arbitrary Waveforms:  
8 to 8,192 points:  
8,193 to 12,287 points:  
12,288 to 16,000 points:  
< 100 kHz:  
100 kHz to 1 MHz:  
1 MHz to 15 MHz:  
1 MHz to 15 MHz:  
1.5% (0.15 dB)  
2% (0.2 dB) Ampl 3Vrms  
3.5% (0.3 dB) Ampl<3Vrms  
100 Hz 5 MHz  
100 Hz 2.5 MHz  
100 Hz 200 kHz  
(3)  
Offset (into 50 ):  
5 Vpk ac + dc  
2% of setting + 2 mV  
(4)  
Accuracy:  
Resolution:  
Accuracy:  
10 Hz or 10 digits  
Output Impedance:  
Resolution:  
50 ohms fixed  
10 ppm in 90 days,  
20 ppm in 1 year,  
18 C 28 C  
3 digits, Amplitude and Offset  
Vpp, Vrms, dBm  
Output Units:  
Isolation:  
Temperature Coefficient:  
Aging:  
< 2 ppm / C  
< 10 ppm / yr  
42 Vpk maximum to earth  
Protection:  
Short-circuit protected  
15 Vpk overdrive < 1 minute  
SINEWAVE SPECTRAL PURITY (into 50 )  
Harmonic Distortion  
DC to 20 kHz:  
-70 dBc  
-60 dBc  
-45 dBc  
-35 dBc  
20 kHz to 100 kHz:  
100 kHz to 1 MHz:  
1 MHz to 15 MHz:  
(1) Add 1/10th of output amplitude and offset specification  
per C for operation outside of 18 C to 28 C range  
(1-year specification).  
Total Harmonic Distortion  
DC to 20 kHz:  
< 0.04%  
(2) 100 mVpp 20 Vpp amplitude into open-circuit load.  
Spurious (non-harmonic)  
Output (DC to 1 MHz):  
Output (> 1 MHz):  
(3) Offset  
2 X peak-to-peak amplitude.  
< -65 dBc  
< -65 dBc + 6 dB/octave  
(4) For square wave outputs, add 2% of output amplitude  
additional error.  
Phase Noise:  
< -55 dBc in a 30 kHz band  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 1 Specifications  
Agilent 33120A Function Generator  
1
SYSTEM CHARACTERISTICS  
MODULATION CHARACTERISTICS  
(2)  
(3)  
AM Modulation  
Configuration Times  
(3)  
Carrier -3 dB Freq:  
Modulation:  
10 MHz (typical)  
Function Change:  
Frequency Change:  
Amplitude Change:  
Offset Change:  
80 ms  
30 ms  
Any internal waveform plus Arb  
10 mHz to 20 kHz ( 0.05% to  
2.5 kHz, then decreases linearly  
to 0.4% at upper limit)  
0% to 120%  
Frequency:  
30 ms  
10 ms  
100 ms  
< 350 ms  
Select User Arb:  
Depth:  
Source:  
Modulation Parameter Change:  
Internal / External  
FM Modulation  
Modulation:  
Any internal waveform plus Arb  
10 mHz to 10 kHz ( 0.05% to  
600 Hz, then decreases linearly  
to 0.8% at upper limit)  
10 mHz to 15 MHz  
Arb Download Times over GPIB:  
Frequency:  
(4)  
Arb Length  
Binary  
ASCII Integer ASCII Real  
Peak Deviation:  
Source:  
Internal Only  
16,000 points  
8,192 points  
4,096 points  
2,048 points  
8 sec  
4 sec  
2.5 sec  
1.5 sec  
81 sec  
42 sec  
21 sec  
11 sec  
100 sec  
51 sec  
26 sec  
13 sec  
Burst Modulation  
Carrier Frequency:  
Count:  
5 MHz max.  
1 to 50,000 cycles, or Infinite  
-360 to +360  
Start Phase:  
(5)  
Arb Download Times over RS-232 at 9600 Baud:  
Internal Rate:  
Gate Source:  
Trigger Source:  
10 mHz to 50 kHz 1%  
Internal or External Gate  
(1)  
(6)  
Single, External, or Internal Rate  
Arb Length  
Binary  
ASCII Integer ASCII Real  
FSK Modulation  
16,000 points  
8,192 points  
4,096 points  
2,048 points  
35 sec  
18 sec  
10 sec  
6 sec  
101 sec  
52 sec  
27 sec  
14 sec  
134 sec  
69 sec  
35 sec  
18 sec  
Frequency Range:  
10 mHz to 15 MHz ( 0.05% to  
600 Hz, then decreases linearly  
to 4% at upper limit)  
Internal Rate:  
Source:  
10 mHz to 50 kHz  
Internal / External (1 MHz max.)  
FREQUENCY SWEEP  
Type:  
Linear or Logarithmic  
Direction:  
Start F / Stop F:  
Time:  
Up or Down  
10 mHz to 15 MHz  
1 ms to 500 sec 0.1%  
Single, External, or Internal  
(1) Trigger source ignored when External Gate is selected.  
(2) Time to change parameter and output the new signal.  
(3) Modulation or sweep off.  
Source:  
REAR-PANEL INPUTS  
External AM  
Modulation:  
5 Vpk = 100% Modulation  
5 k Input Resistance  
(4) Times for 5-digit and 12-digit numbers.  
(5) For 4800 baud, multiply the download times by two;  
External Trigger/FSK  
For 2400 baud, multiply the download times by four, etc.  
(1)  
Burst Gate:  
Latency:  
Jitter:  
TTL (low true)  
1.3 s  
25 ns  
(6) Time for 5-digit numbers. For 12-digit numbers,  
multiply the 5-digit numbers by two.  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 1 Specifications  
Agilent 33120A Function Generator  
GENERAL SPECIFICATIONS  
(1)  
Power Supply:  
100V / 120V / 220V / 240V 10%  
(switch selectable)  
Safety Designed to:  
EMC:  
EN61010, CSA1010, UL-1244  
EN61326, 1:1997 + 1A:1998  
Power-Line Frequency:  
50 Hz to 60 Hz 10% and  
400 Hz 10%. Automatically  
sensed at power-on.  
Vibration and Shock:  
MIL-T-28800E, Type III, Class 5  
(data on file)  
Power Installation:  
CAT II  
Acoustic Noise:  
Warm-Up Time:  
30 dBa  
Power Consumption:  
50 VA peak (28 W average)  
1 hour  
Operating Environment: 0 C to 55 C  
80% Relative Humidity to 40 C  
Warranty:  
3 years standard  
IEEE-488 and RS-232 standard  
SCPI-1993, IEEE-488.2  
Indoor or sheltered location  
Remote Interface:  
Programming Language:  
Accessories Included:  
Storage Environment:  
State Storage Memory:  
-40 C to 70 C  
Power-off state automatically  
saved. Three (3) User-  
Configurable Stored States,  
Arbitrary waveforms stored  
separately.  
User’s Guide, Service Guide,  
Quick Reference Card,  
IntuiLink Arb software,  
RS-232 cable, Test Report,  
and power cord.  
Dimensions (W x H x D)  
Bench Top:  
Rack Mount:  
254.4 mm x 103.6 mm x 374 mm  
212.6 mm x 88.5 mm x 348.3 mm  
N10149  
Weight:  
4 kg (8.8 lbs)  
(1) For 400 Hz operation at 120 Vac, use the 100 Vac  
line-voltage setting.  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 1 Specifications  
Agilent 33120A Function Generator  
1
PRODUCT DIMENSIONS  
TOP  
All dimensions are  
shown in millimeters.  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2
2
Quick Start  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Quick Start  
One of the first things you will want to do with your function generator is  
to become acquainted with its front panel. We have written the exercises  
in this chapter to prepare the function generator for use and help you get  
familiar with some of the front-panel operations.  
The front panel has two rows of keys to select various functions and  
operations. Most keys have a shifted function printed in blue above the  
key. To perform a shifted function, press Shift (the Shift annunciator will  
turn on). Then, press the key that has the desired label above it. For  
example, to select the AM (amplitude modulation) function, press  
Shift AM (the shifted version of the  
key).  
If you accidentally press Shift , just press it again to turn off the Shift  
annunciator.  
Most keys also have a number printed in green next to the key. To enable  
the number mode, press Enter Number (the Num annunciator will turn on).  
Then, press the keys that have the desired numbers printed next to them.  
For example, to select the number “10”, press Enter Number  
(next to the and Recall keys).  
1
0
If you accidentally press Enter Number , just press Shift Cancel to turn  
off the Num annunciator.  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To prepare the function generator for use  
To prepare the function generator for use  
The following steps help you verify that the function generator is  
ready for use.  
2
1 Check the list of supplied items.  
Verify that you have received the following items with your function  
generator. If anything is missing, contact your nearest Agilent  
Technologies Sales Office.  
One power cord.  
One RS-232 serial cable.  
One User’s Guide.  
This Service Guide.  
á
One folded Quick Reference card.  
Certificate of Calibration.  
Agilent IntuiLink Arb Waveform Generation Software.  
2 Connect the power cord and turn on the function generator.  
If the function generator does not turn on, see chapter 6 for troubleshooting  
information. The front-panel display will light up while the function  
generator performs its power-on self-test. The GPIB bus address is  
displayed. Notice that the function generator powers up in the sine wave  
function at 1 kHz with an amplitude of 100 mV peak-to-peak (into a 50W  
termination).  
To review the power-on display with all annunciators turned on,  
hold down Shift as you turn on the function generator.  
3 Perform a complete self test.  
The complete self-test performs a more extensive series of tests than  
those performed at power-on. Hold down Shift as you press the Power  
switch to turn on the function generator; hold down the key for more than  
5 seconds. The self-test will begin when you release the key.  
If the self-test is successful, “PASS” is displayed on the front panel.  
If the self-test is not successful, “FAIL” is displayed and the ERROR  
annunciator turns on. See chapter 6 for instructions on returning the  
function generator to Agilent for service.  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
If the function generator does not turn on  
If the function generator does not turn on  
Use the following steps to help solve problems you might experience  
when turning on the function generator. If you need more help,  
see chapter 6 for instructions on returning the function generator to  
Agilent for service.  
1 Verify that there is ac power to the function generator.  
First, verify that the function generator’s Power switch is in the  
“On” position. Also, make sure that the power cord is firmly plugged into  
to the power module on the rear panel. You should also make sure that  
the power source you plugged the function generator into is energized.  
2 Verify the power-line voltage setting.  
The line voltage is set to the proper value for your country when the  
function generator is shipped from the factory. Change the voltage  
setting if it is not correct. The settings are: 100, 120, 220, or 240 Vac  
(for 230 Vac operation, use the 220 Vac setting).  
See the next page if you need to change the line-voltage setting.  
3 Verify that the power-line fuse is good.  
The function generator is shipped from the factory with a 500 mAT fuse  
installed. This is the correct fuse for all line voltages.  
See the next page if you need to change the power-line fuse.  
To replace the 500 mAT fuse, order Agilent part number 2110-0458.  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
If the function generator does not turn on  
2 Remove the line-voltage selector from  
1 Remove the power cord. Remove the  
the assembly.  
fuse-holder assembly from the rear panel.  
2
Fuse: 500 mAT (for all line voltages)  
Part Number: 2110-0458  
3 Rotate the line-voltage selector until the  
4 Replace the fuse-holder assembly in  
correct voltage appears in the window.  
the rear panel.  
100, 120, 220 (230), or 240 Vac  
Verify that the correct line voltage is selected and the power-line fuse is good.  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To adjust the carrying handle  
To adjust the carrying handle  
To adjust the position, grasp the handle by the sides and pull outward.  
Then, rotate the handle to the desired position.  
Bench-top viewing positions  
Carrying position  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To set the output frequency  
To set the output frequency  
At power-on, the function generator outputs a sine wave at 1 kHz with  
an amplitude of 100 mV peak-to-peak (into a 50W termination).  
The following steps show you how to change the frequency to 1.2 MHz.  
2
Freq  
1 Enable the frequency modify mode.  
The displayed frequency is either the power-on value or the previous  
frequency selected. When you change functions, the same frequency is  
used if the present value is valid for the new function.  
1.000,000,0 KHz  
1
Enter Number  
2 Enter the magnitude of the desired frequency.  
1
.
2
Notice that the Num annunciator turns on and “ENTER NUM” flashes on  
the display, indicating that the number mode is enabled.  
1.2  
To cancel the number mode, press Shift Cancel .  
MHz  
m Vpp  
3 Set the units to the desired value.  
The units are selected using the arrow keys on the right side of the front  
panel. As soon as you select the units, the function generator outputs the  
waveform with the displayed frequency. To turn off the flashing digit,  
move the cursor to the left of the display using the arrow keys.  
1.200,000,0 MHz  
1
You can also use the knob and arrow keys to enter a number.  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To set the output amplitude  
To set the output amplitude  
At power-on, the function generator outputs a sine wave with an  
amplitude of 100 mV peak-to-peak (into a 50W termination).  
The following steps show you how to change the amplitude to 50 mVrms.  
Ampl  
1 Enable the amplitude modify mode.  
The displayed amplitude is either the power-on value or the previous  
amplitude selected. When you change functions, the same amplitude is  
used if the present value is valid for the new function.  
100.0 mVPP  
1
Enter Number  
2 Enter the magnitude of the desired amplitude.  
5
0
Notice that the Num annunciator turns on and “ENTER NUM” flashes on  
the display, indicating that the number mode is enabled.  
50  
To cancel the number mode, press Shift Cancel .  
Shift  
3 Set the units to the desired value.  
kHz  
m Vrms  
The units are selected using the arrow keys on the right side of the front  
panel. As soon as you select the units, the function generator outputs the  
waveform with the displayed amplitude. To turn off the flashing digit,  
move the cursor to the left of the display using the arrow keys.  
50.00 mVRMS  
1
You can also use the knob and arrow keys to enter a number.  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To set a dc offset voltage  
To set a dc offset voltage  
At power-on, the function generator outputs a sine wave with a dc offset  
voltage of 0 volts (into a 50W termination). The following steps show you  
how to change the offset to 1.5 mVdc.  
2
Offset  
1 Enable the offset modify mode.  
The displayed offset voltage is either the power-on value or the previous  
offset selected. When you change functions, the same offset is used if the  
present value is valid for the new function.  
+0.000 VDC  
1
Enter Number  
2 Enter the magnitude of the desired offset.  
1
.
5
Notice that the Num annunciator turns on and “ENTER NUM” flashes on  
the display, indicating that the number mode is enabled. Notice that  
toggles the displayed value between + and .  
-1.5  
To cancel the number mode, press Shift Cancel .  
Shift  
3 Set the units to the desired value.  
kHz  
m Vrms  
At this point, the function generator outputs the waveform with the  
displayed offset. Notice that the Offset annunciator turns on, indicating  
that the waveform is being output with an offset. The annunciator will  
turn on when the offset is any value other than 0 volts. To turn off the  
flashing digit, move the cursor to the left of the display using the arrow keys.  
-01.50 mVDC  
1
You can also use the knob and arrow keys to enter a number.  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To set the duty cycle  
To set the duty cycle  
Applies only to square waves. At power-on, the duty cycle for square waves  
is 50%. You can adjust the duty cycle for a square waveform from 20%  
to 80%, in increments of 1% (for frequencies above 5 MHz, the range is 40%  
to 60%). The following steps show you how to change the duty cycle to 45%.  
1 Select the square wave function.  
Notice that the  
annunciator turns on, indicating that the square  
wave function is enabled.  
Shift  
% Duty  
2 Enable the duty cycle modify mode.  
The displayed duty cycle is either the power-on value or the previous  
value selected.  
50 % DUTY  
This message appears on the display for approximately 10 seconds.  
Repeat this step as needed.  
1
Enter Number  
3 Enter the desired duty cycle.  
4
5
Notice that the Num annunciator turns on and “ENTER NUM” flashes on  
the display, indicating that the number mode is enabled.  
45  
To cancel the number mode, press Shift Cancel .  
Enter  
4 Output the waveform with the displayed duty cycle.  
45 % DUTY  
1
You can also use the knob and arrow keys to enter a number.  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To output a stored arbitrary waveform  
To output a stored arbitrary waveform  
There are five built-in arbitrary waveforms stored in non-volatile memory  
for your use. You can output these waveforms directly from non-volatile  
memory. The following steps show you how to output an “exponential rise”  
waveform from memory.  
2
Shift  
Arb List  
1 Display the list of arbitrary waveforms.  
The list contains the five built-in arbitrary waveforms (sinc, negative  
ramp, exponential rise, exponential fall, and cardiac). The list may also  
contain up to four user-defined arbitrary waveform names. The first  
choice on this level is “SINC”.  
SINC  
This message appears on the display for approximately 10 seconds.  
Repeat this step as needed.  
1
>
>
2 Move across to the EXP_RISE choice.  
EXP_RISE  
Enter  
3 Select and output the displayed arbitrary waveform.  
Notice that the Arb annunciator turns on, indicating that the output is an  
arbitrary waveform. The waveform is output using the present settings  
for frequency, amplitude, and offset unless you change them.  
The selected waveform is now assigned to the Arb key. Whenever you  
press this key, the selected arbitrary waveform is output.  
1
You can also use the knob to scroll left or right through the choices in the list.  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To output a dc voltage  
To output a dc voltage  
In addition to generating waveforms, you can also output a dc voltage in  
the range 5 Vdc (into a 50W termination). The following steps show you  
how to output +155 mVdc.  
Offset  
1 Press the  
key and hold it down for more than 2 seconds.  
To enter the dc voltage mode, press the Offset key or any key in the top row  
of function keys and hold it down for more than 2 seconds. The displayed  
voltage is either the power-on value or the previous offset voltage selected.  
DCV  
+0.000 VDC  
1
Enter Number  
2 Enter the magnitude of the desired voltage.  
1
5
5
Notice that the Num annunciator turns on and “ENTER NUM” flashes on  
the display, indicating that the number mode is enabled.  
155  
To cancel the number mode, press Shift Cancel .  
Shift  
3 Set the units to the desired value.  
kHz  
m Vrms  
At this point, the function generator outputs the displayed dc voltage.  
Notice that the Offset annunciator turns on (all other annunciators are  
off), indicating that a dc voltage is being output. The annunciator will  
turn on when the offset is any value other than 0 volts.  
+155.0 mVDC  
1
You can also use the knob and arrow keys to enter a number.  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To store the instrument state  
To store the instrument state  
You can store up to three different instrument states in non-volatile  
memory. This enables you to recall the entire instrument configuration  
with just a few key presses from the front panel. The following steps show  
you how to store and recall a state.  
2
1 Set up the function generator to the desired configuration.  
The state storage feature “remembers” the function, frequency,  
amplitude, dc offset, duty cycle, as well as any modulation parameters.  
Shift  
Store  
2 Turn on the state storage mode.  
Three memory locations (numbered 1, 2, and 3) are available to store  
instrument configurations. The instrument configuration is stored in  
non-volatile memory and is remembered when power has been off.  
STORE 1  
This message appears on the display for approximately 10 seconds.  
Repeat this step as needed.  
1
3 Store the instrument state in memory location “2”.  
Use the up and down arrow keys to select the memory location.  
STORE 2  
To cancel the store operation, press Shift Store again or let the display  
time-out after 10 seconds.  
Enter  
4 Save the instrument state.  
The instrument state is now stored. To recall the stored state, turn to the  
next page.  
1
You can also use the knob or “enter number” mode to enter a memory location.  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To store the instrument state  
To verify that the state was stored properly, you can turn the power off  
before recalling the state.  
Recall  
5 Recall the stored instrument state.  
To recall the stored state, you must use the same memory location used  
previously to store the state. Use the up and down arrow keys to change  
the displayed storage location.  
RECALL 2  
To cancel the restore operation, press Recall again.  
This message appears on the display for approximately 10 seconds.  
Repeat this step as needed.  
Enter  
6 Restore the instrument state.  
The function generator should now be configured in the same state as  
when you stored the setup on the previous page.  
When power is turned off, the function generator automatically stores  
its state in memory location “0”. You can recall the power-down state,  
but you cannot store the state to location “0” from the front panel.  
Use the POWER ON ENABLE command in the SYS MENU to  
automatically recall the power-down state when power is turned on.  
See chapter 3 for more information on using the front-panel menus.  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To rack mount the function generator  
To rack mount the function generator  
You can mount the function generator in a standard 19-inch rack  
cabinet using one of three optional kits available. Instructions and  
mounting hardware are included with each rack-mounting kit.  
Any Agilent System II instrument of the same size can be rack-mounted  
beside the 33120A Function Generator.  
2
Remove the carrying handle, and the front and rear rubber bumpers,  
before rack-mounting the function generator.  
To remove the handle, rotate it to the vertical position and pull the ends outward.  
Front  
Rear (bottom view)  
To remove the rubber bumper, stretch a corner and then slide it off.  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2 Quick Start  
To rack mount the function generator  
To rack mount a single instrument, order adapter kit 5063-9240.  
To rack mount two instruments side-by-side, order lock-link kit 5061-9694  
and flange kit 5063-9212.  
To install one or two instruments in a sliding support shelf, order shelf 5063-9255,  
and slide kit 1494-0015 (for a single instrument, also order filler panel 5002-3999).  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3
3
Front-Panel  
Menu Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Front-Panel Menu Operation  
By now you should be familiar with some of the basic features of the front  
panel. Chapter 2 shows you how to prepare the function generator for use  
and describes a few of the front-panel features. If you are not familiar  
with this information, we recommend that you read chapter 2, “Quick Start,”  
starting on page 19.  
Chapter 3 introduces you to the use of the front-panel menu. This chapter  
does not give a detailed description of every front-panel key or menu  
operation. It does, however, give you an overview of front-panel menu  
operations related to verification, adjustment and service. See chapter 3  
“Features and Functions” in the User’s Guide for a complete discussion of  
the function generator’s capabilities and operation.  
If you purchased the Phase-Lock Option for the 33120A, an additional  
menu (G: PHASE MENU) is available from the front panel. For inform-  
ation on using the Phase-Lock Option, refer to the User’s and Service  
Guide included with Option 001.  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
Front-panel menu reference  
Front-panel menu reference  
A: MODulation MENU  
1: AM SHAPE Õ 2: AM SOURCE Õ 3: FM SHAPE Õ 4: BURST CNT Õ 5: BURST RATE Õ  
Õ 6: BURST PHAS Õ 7: BURST SRC Õ 8: FSK FREQ Õ 9: FSK RATE Õ 10: FSK SRC  
1: AM SHAPE  
2: AM SOURCE  
3: FM SHAPE  
4: BURST CNT  
5: BURST RATE  
6: BURST PHAS  
7: BURST SRQ  
8: FSK FREQ  
9: FSK RATE  
Selects the shape of the AM modulating waveform.  
Enables or disables the internal AM modulating source.  
Selects the shape of the FM modulating waveform.  
Sets the number of cycles per burst (1 to 50,000 cycles).  
Sets the burst rate in Hz for an internal burst source.  
Sets the starting phase angle of a burst (-360 to +360 degrees).  
Selects an internal or external gate source for burst modulation.  
Sets the FSK “hop” frequency.  
3
Selects the internal FSK rate between the carrier and FSK frequency.  
Selects an internal or external source for the FSK rate.  
10: FSK SRC  
B: SWP (Sweep) MENU  
1: START F Õ 2: STOP F Õ 3: SWP TIME Õ 4: SWP MODE  
1: START F  
2: STOP F  
3: SWP TIME  
4: SWP MODE  
Sets the start frequency in Hz for sweeping.  
Sets the stop frequency in Hz for sweeping.  
Sets the repetition rate in seconds for sweeping.  
Selects linear or logarithmic sweeping.  
C: EDIT MENU *  
1: NEW ARB Õ 2: POINTS Õ [3: LINE EDIT] Õ [4: POINT EDIT] Õ [5: INVERT] Õ [6: SAVE AS] Õ 7:DELETE  
1: NEW ARB  
2: POINTS  
3: LINE EDIT  
4: POINT EDIT  
5: INVERT  
6: SAVE AS  
7: DELETE  
Initiates a new arb waveform or loads the selected arb waveform.  
Sets the number of points in a new arb waveform (8 to 16,000 points).  
Performs a linear interpolation between two points in the arb waveform.  
Edits the individual points of the selected arb waveform.  
Inverts the selected arb waveform by changing the sign of each point.  
Saves the current arb waveform in non-volatile memory.  
Deletes the selected arb waveform from non-volatile memory.  
* The commands enclosed in square brackets ( [ ] ) are “hidden” until you make a selection from the NEW ARB  
command to initiate a new edit session.  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
Front-panel menu reference  
D: SYStem MENU  
1: OUT TERM Õ 2: POWER ON Õ 3: ERROR Õ 4: TEST Õ 5: COMMA Õ 6:REVISION  
1: OUT TERM  
2: POWER ON  
3: ERROR  
Selects the output termination (50 or high impedance).  
Enables or disables automatic power-up in power-down state “0”.  
Retrieves errors from the error queue (up to 20 errors).  
Performs a complete self-test.  
4: TEST  
5: COMMA  
6: REVISION  
Enables or disables a comma separator between digits on the display.  
Displays the function generator’s firmware revision codes.  
E: Input / Output MENU  
1: HPIB ADDR Õ 2: INTERFACE Õ 3: BAUD RATE Õ 4: PARITY Õ 5: LANGUAGE  
1: HPIB ADDR  
2: INTERFACE  
3: BAUD RATE  
4: PARITY  
Sets the GPIB bus address (0 to 30).  
Selects the GPIB or RS-232 interface.  
Selects the baud rate for RS-232 operation.  
Selects even, odd, or no parity for RS-232 operation.  
Verifies the interface language: SCPI.  
5: LANGUAGE  
F: CALibration MENU *  
1: SECURED Õ [1: UNSECURED] Õ [2: CALIBRATE] Õ 3: CAL COUNT Õ 4: MESSAGE  
1: SECURED  
The function generator is secured against calibration; enter code to unsecure.  
The function generator is unsecured for calibration; enter code to secure.  
Performs individual calibrations; must be UNSECURED.  
Reads the total number of times the function generator has been calibrated.  
Reads the calibration string (up to 11 characters) entered from remote.  
1: UNSECURED  
2: CALIBRATE  
3: CAL COUNT  
4: MESSAGE  
* The commands enclosed in square brackets ( [ ] ) are “hidden” unless the function generator is UNSECURED for  
calibration.  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
A front-panel menu tutorial  
A front-panel menu tutorial  
This section is a step-by-step tutorial which shows you how to use the  
front-panel menu. We recommend that you spend a few minutes with this  
tutorial to get comfortable with the structure and operation of the menu  
before attempting verification, calibration, or adjustments.  
The menu is organized in a top-down tree structure with three levels  
(menus, commands, and parameters). You move down  
or up  
¾
the menu tree to get from one level to the next. Each of the three levels  
has several horizontal choices which you can view by moving left  
or right > .  
<
3
Menus  
Commands  
Parameters  
The menu is organized in a top-down tree structure with three levels.  
œ To turn on the menu, press Shift Menu On/Off .  
œ To turn off the menu, press Shift Menu On/Off .  
œ To execute a menu command, press Enter .  
œ To recall the last menu command that was executed,  
press Shift Recall Menu .  
œ To turn off the menu at any time without saving changes,  
press Shift Cancel .  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
A front-panel menu tutorial  
Messages Displayed During Menu Use  
TOP OF MENU You pressed  
¾
while on the “MENUS” level; this is the top  
level of the menu and you cannot go any higher.  
To turn off the menu, press Shift Menu On/Off . To move across the choices on  
a level, press  
<
or  
>
. To move down a level, press  
¿
.
MENUS You are on the “MENUS” level. Press < or  
>
to view the choices.  
COMMANDS You are on the “COMMANDS” level. Press < or > to view the  
command choices within the selected menu group.  
PARAMETER You are on the “PARAMETER” level. Press < or > to view  
and edit the parameter for the selected command.  
MENU BOTTOM You pressed  
¿
while on the “PARAMETER” level; this is the  
bottom level of the menu and you cannot go any lower.  
To turn off the menu, press Shift Menu On/Off . To move up a level, press  
¾
.
ENTERED The change made on the “PARAMETER” level is saved. This is  
displayed after you press Enter (Menu Enter) to execute the command.  
MIN VALUE The value you specified on the “PARAMETER” level is too small for  
the selected command. The minimum value allowed is displayed for you to edit.  
MAX VALUE The value you specified on the “PARAMETER” level is too large for  
the selected command. The maximum value allowed is displayed for you to edit.  
EXITING You will see this message if you turn off the menu by pressing  
Shift Menu On/Off or Shift Cancel . You did not edit any values on the  
PARAMETER” level and changes were NOT saved.  
NOT ENTERED You will see this message if you turn off the menu by pressing  
Shift Menu On/Off or Shift Cancel . You did some editing of parameters but  
the changes were NOT saved. Press Enter (Menu Enter) to save changes  
made on the “PARAMETER” level.  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
A front-panel menu tutorial  
Menu Example 1  
The following steps show you how to turn on the menu, move up and  
down between levels, move across the choices on each level, and turn off  
the menu. In this example, you will restore the function generator to the  
power-on default state. This procedure is recommended before performing  
the verification procedures in chapter 4.  
Shift  
1 Turn on the menu.  
Menu On/Off  
You enter the menu on the “MENUS” level. The MOD MENU is your first  
choice on this level.  
3
A: MOD MENU  
1
>
>
>
2 Move across to the SYS MENU choice on this level.  
There are six menu group choices available on the “MENUS” level. Each  
choice has a letter prefix for easy identification (A: , B: , etc.).  
D: SYS MENU  
3 Move down to the “COMMANDS” level within the SYS MENU.  
The OUT TERM command is your first choice on this level.  
1: OUT TERM  
1
>
4 Move across to the POWER ON command on this level.  
There are six command choices available in the SYS MENU. Each choice  
on this level has a number prefix for easy identification (1: , 2: , etc.).  
2: POWER ON  
1
You can also use the knob to scroll left or right through the choices on each  
level of the menu.  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
A front-panel menu tutorial  
5 Move down a level to the “PARAMETER” choices.  
The first parameter choice is “DEFAULT” for the POWER ON command  
(“DEFAULT” is the factory setting and is stored in non-volatile memory).  
DEFAULT  
1
>
6 Move across to the “LAST STATE” choice.  
There are two parameter choices for POWER ON.  
LAST STATE  
Enter  
7 Save the change and turn off the menu.  
The function generator beeps and displays a message to show that the  
change is now in effect. You are then exited from the menu.  
ENTERED  
8 Cycle the power to restore the default values.  
Turn the function generator OFF and then ON. The default output  
state will now be in effect (1 kHz sine wave, 100 mV peak-to-peak,  
50W termination).  
1
You can also use the knob to scroll left or right through the choices on each  
level of the menu.  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
A front-panel menu tutorial  
Menu Example 2  
Some commands in the menu require that you enter a numeric  
parameter value. The following steps show you how to enter a number in  
the menu. For this example, you will change the output amplitude.  
Ampl  
1 Select amplitude adjustment  
The function generator displays the current output amplitude.  
100.0 mVPP  
3
<
2 Move the flashing cursor over to edit the first digit.  
The cursor movement wraps around.  
100.0 mVPP  
1
^
^
^
3 Increment the first digit until 300.0 mVPP is displayed.  
The output amplitude of the function changes as you adjust the  
displayed value.  
300.0 mVPP  
1
You can also use the knob and arrow keys to enter a number.  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
To select the output termination  
To select the output termination  
The function generator has a fixed output impedance of 50 ohms on the  
OUTPUT terminal. You can specify whether you are terminating the  
output into a 50W load or an open circuit. Incorrect impedance matching  
between the source and load will result in an output amplitude or dc offset  
which does not match the specified value.  
Shift  
1 Turn on the menu.  
Menu On/Off  
A: MOD MENU  
1
>
>
>
2 Move across to the SYS MENU choice on this level.  
D: SYS MENU  
3 Move down a level to the OUT TERM command.  
1: OUT TERM  
1
>
4
Move down a level and then across to the HIGH Z choice.  
With the output termination set to “HIGH Z”, the function generator  
allows you to set the unloaded (open circuit) output voltage.  
HIGH Z  
Enter  
5 Save the change and turn off the menu.  
The function generator beeps and displays a message to show that the  
change is now in effect. You are then exited from the menu.  
1
You can also use the knob to scroll left or right through the choices on each  
level of the menu.  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
To output a modulated waveform  
To output a modulated waveform  
A modulated waveform consists of a carrier and a modulating waveform.  
In AM (amplitude modulation), the amplitude of the carrier is varied by  
the amplitude of the modulating waveform. For this example, you will  
output an AM waveform with 80% modulation depth. The carrier will be a  
5 kHz sine wave and the modulating waveform will be a 200 Hz sine wave.  
3
1 Select the function, frequency, and amplitude of the carrier.  
For the carrier waveform, you can select a sine, square, triangle, ramp,  
or arbitrary waveform. For this example, select a 5 kHz sine wave with  
an amplitude of 5 Vpp.  
Shift  
AM  
2 Select AM.  
Notice that the AM annunciator turns on.  
Shift  
<
3 Use the menu to select the shape of the modulating waveform.  
Recall Menu  
After you enable the AM function, the “recall menu” key will  
automatically take you to the AM SHAPE command in the MOD MENU.  
1: AM SHAPE  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
To output a modulated waveform  
4 Move down a level verify that “SINE” is selected.  
For the modulating waveform, you can select a sine, square, triangle,  
ramp, noise, or arbitrary waveform. For this example, you will modulate  
the carrier with a sine waveform. Notice that the AM annunciator flashes,  
indicating that the displayed parameter is for AM.  
SINE  
Enter  
5 Save the change and turn off the menu.  
The modulating waveform is now a sine waveform.  
ENTERED  
Shift  
Freq  
6 Set the modulating frequency to 200 Hz.  
Notice that the AM annunciator flashes, indicating that the displayed  
frequency is the modulating frequency. Also notice that the modulating  
frequency is displayed with fewer digits than the carrier frequency.  
MOD 200.0 Hz  
This message appears on the display for approximately 10 seconds.  
Repeat this step as needed.  
Shift  
Level  
7 Set the modulation depth to 80%.  
Notice that the AM annunciator flashes, indicating that the displayed  
percentage is the AM depth (also called percent modulation).  
080 % DEPTH  
This message appears on the display for approximately 10 seconds.  
Repeat this step as needed.  
At this point, the function generator outputs the AM waveform with the  
specified modulation parameters.  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
To unsecure the function generator for calibration  
To unsecure the function generator for calibration  
The function generator can use a calibration security code to prevent  
unauthorized or accidental calibration. This procedure shows you how to  
unsecure the function generator for calibration.  
Shift  
1 Turn on the menu.  
Menu On/Off  
<
A: MOD MENU  
3
2 Move across to the CAL MENU choice on this level.  
F: CAL MENU  
3 Move down a level to the SECURED command.  
1: SECURED  
If the display shows UNSECURED, you do not need to perform this  
procedure to execute a calibration.  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3 Front-Panel Menu Operation  
To unsecure the function generator for calibration  
4 Move down to the “parameters” level.  
^000000:CODE  
0
1
3
2
3
0
5 Unsecure the function generator by entering the security code.  
ENTER  
^033120:CODE  
The security code is set to “HP33120” when the function generator is  
shipped from the factory. The security code is stored in non-volatile  
memory and does not change when the power has been off or after a  
remote interface reset.  
To enter the security code from the front panel, enter only the six digits.  
To enter the security code from the remote interface, you may enter up to  
12 characters. Use the knob or arrow keys to move left or right between  
digits. Use the up or down arrow keys to change the digits.  
To re-secure the function generator following a calibration, perform this  
procedure again.  
Additional information about the calibration security feature is given  
on page 64.  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4
4
Calibration  
Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Calibration Procedures  
This chapter contains procedures for verification of the function  
generator’s performance and adjustment (calibration). The chapter is  
divided into the following sections:  
œ Agilent Calibration Services . . . . . . . . . . . . . . 51  
œ Calibration Interval . . . . . . . . . . . . . . . . . . . 51  
œ Time Required for Calibration . . . . . . . . . . . . . 51  
œ Automating Calibration Procedures . . . . . . . . . . 52  
œ Recommended Test Equipment . . . . . . . . . . . . . 52  
œ Test Considerations . . . . . . . . . . . . . . . . . . . 53  
œ Performance Verification Tests . . . . . . . . . . . . . 54  
œ Frequency Verification . . . . . . . . . . . . . . . . . 56  
œ Function Gain and Linearity Verification . . . . . . . 56  
œ DC Function Offset Verification . . . . . . . . . . . . 57  
œ AC Amplitude Verification . . . . . . . . . . . . . . . 57  
œ Amplitude Flatness Verification . . . . . . . . . . . . 60  
œ AM Modulation Depth Verification . . . . . . . . . . . 61  
œ Optional Performance Verification Tests . . . . . . . . 62  
œ Calibration Security Code . . . . . . . . . . . . . . . . 64  
œ Calibration Count . . . . . . . . . . . . . . . . . . . . 66  
œ Calibration Message . . . . . . . . . . . . . . . . . . . 66  
œ General Calibration/Adjustment Procedure . . . . . . 67  
œ Aborting a Calibration in Progress . . . . . . . . . . . 69  
œ Frequency and Burst Rate Adjustment . . . . . . . . 69  
œ Function Gain and Linearity Adjustment . . . . . . . 70  
œ AC Amplitude Adjustment (High-Z) . . . . . . . . . . 70  
œ Modulation Adjustment . . . . . . . . . . . . . . . . . 72  
œ AC Amplitude Adjustment (50W) . . . . . . . . . . . . 73  
œ DC Output Adjustment . . . . . . . . . . . . . . . . . 76  
œ Duty Cycle Adjustment . . . . . . . . . . . . . . . . . 77  
œ AC Amplitude Flatness Adjustment . . . . . . . . . . 77  
œ Output Amplifier Adjustment (Optional) . . . . . . . 80  
œ Error Messages . . . . . . . . . . . . . . . . . . . . . 81  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Agilent Calibration Services  
Closed-Case Electronic Calibration The function generator features  
closed-case electronic calibration since no internal mechanical  
adjustments are required for normal calibration. The function generator  
calculates correction factors based upon the input reference value you set.  
The new correction factors are stored in non-volatile memory until the  
next calibration adjustment is performed (non-volatile memory does not  
change when power has been off or after a remote interface reset).  
Agilent Calibration Services  
When your function generator is due for calibration, contact your local  
Agilent Service Center for a low-cost recalibration. The 33120A Function  
Generator is supported on automated calibration systems which allow  
Agilent to provide this service at competitive prices. Calibrations to  
MIL-STD-45662 are also available at competitive prices.  
4
Calibration Interval  
The function generator should be calibrated on a regular interval  
determined by the measurement accuracy requirements of your  
application. A 1- or 2-year interval is adequate for most applications.  
Agilent does not recommend extending calibration intervals beyond  
two years for any application.  
Whatever calibration interval you select, Agilent recommends that complete  
re-adjustment should always be performed at the calibration interval.  
This will increase your confidence that the 33120A will remain within  
specification for the next calibration interval. This criteria for re-adjustment  
provides the best long-term stability. Performance data measured using this  
method can be used to extend future calibration intervals.  
Time Required for Calibration  
The 33120A can be automatically calibrated under computer control.  
With computer control you can perform the complete calibration  
procedure and performance verification tests in less than 15 minutes.  
Manual calibrations using the recommended test equipment will take  
approximately 45 minutes.  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Automating Calibration Procedures  
Automating Calibration Procedures  
You can automate the complete verification and adjustment procedures  
outlined in this chapter if you have access to programmable test  
equipment. You can program the instrument configurations specified for  
each test over the remote interface. You can then enter readback  
verification data into a test program and compare the results to the  
appropriate test limit values.  
You can also enter calibration constants from the remote interface.  
Remote operation is similar to the local front-panel procedure. You can  
use a computer to perform the adjustment by first selecting the required  
setup. The calibration value is sent to the function generator and then  
the calibration is initiated over the remote interface. The function  
generator must be unsecured prior to initiating the calibration procedure.  
For further detailing on programming the function generator, see  
chapters 3 and 4 in the Agilent 33120A User’s Guide.  
Recommended Test Equipment  
The test equipment recommended for the performance verification and  
adjustment procedures is listed below. If the exact instrument is not  
available, use the accuracy requirements shown to select substitute  
calibration standards.  
Instrument  
Requirements  
Recommended Model  
Use*  
Q,P,O,T  
50 W feedthrough load  
50 W – 0.1 W  
6 1/2 digit Digital  
Multimeter (DMM)  
20 Vdc 0.01%  
Integrating ACrms  
10 Vacrms 0.1%  
Agilent 34401A  
3 Volt  
Q,P,T  
Q,P  
Thermal Voltage Converter  
1kHz to 15 MHz  
(50 W termination type)  
Agilent E4418A with  
Agilent 8482A  
and 20 dB attenuator  
or  
100 kHz to 15 MHz  
Power Meter  
or  
Wideband ACrms Meter  
1 VAC rms 0.5%  
1 kHz to 20 MHz  
Frequency Meter  
Oscilloscope  
1 ppm accuracy  
100 MHz  
Agilent 53131A  
Agilent 54624A  
Agilent 8560EC  
Q,P,T  
T
Spectrum Analyzer  
Response to 90 MHz  
O
*
Q = Quick Verification  
P = Performance Verification Tests  
O= Optional Verification Tests  
T = Troubleshooting  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Test Considerations  
Test Considerations  
To ensure proper instrument operation, verify that you have selected the  
correct power line voltage prior to attempting any test procedure in this  
chapter. See page 22 in chapter 2 for more information.  
For optimum performance, all test procedures should comply with the  
following recommendations:  
œ Verify the function generator is set to the default power on state  
(power on default). A procedure is given on page 41.  
œ Make sure that the calibration ambient temperature is stable and  
between 18 C and 28 C.  
œ Make sure ambient relative humidity is less than 80%.  
œ Allow a 1-hour warm-up period before verification or adjustment.  
œ Use only RG-58 or equivalent 50W cable.  
4
œ Keep cables as short as possible, consistent with the impedance  
requirements.  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Performance Verification Tests  
Performance Verification Tests  
The performance verification tests use the function generator’s  
specifications listed in chapter 1, “Specifications,” starting on page 13.  
You can perform four different levels of performance verification tests:  
œ Self-Test A series of internal verification tests that give a high  
confidence that the function generator is operational.  
Q
P
œ Quick Verification A combination of the internal self-tests and  
selected verification tests.  
œ Performance Verification Tests An extensive set of tests that are  
recommended as an acceptance test when you first receive the function  
generator or after performing adjustments.  
œ Optional Verification Tests Tests not performed with every  
calibration. These tests can can be used to verify additional instrument  
specifications following repairs to specific circuits.  
O
Self-Test  
A brief power-on self-test occurs automatically whenever you turn on the  
function generator. This limited test assures that the function generator  
is capable of operation.  
To perform a complete self-test hold down the Shift key as you press the  
Power switch to turn on the function generator; hold down the key for  
more than 5 seconds (a complete description of these tests can be found in  
chapter 6). The function generator will automatically perform the  
complete self-test procedure when you release the key. The self-test will  
complete in approximately 5 seconds.  
You can perform many tests individually (or all tests at once) using the  
TEST command in the SYS MENU. You can also perform a self-test from  
the remote interface (see chapter 3 in the Agilent 33120A User’s Guide).  
œ If the self-test is successful, “PASS” is displayed on the front panel.  
œ If the self-test fails, “FAIL” is displayed and the ERROR annunciator  
turns on. If repair is required, see chapter 6, “Service,” for further details.  
œ If all tests pass, you have a high confidence (90%) that the function  
generator is operational.  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Performance Verification Tests  
Quick Performance Check  
The quick performance check is a combination of internal self-test and an  
abbreviated performance test (specified by the letter Q in the performance  
verification tests). This test provides a simple method to achieve high  
confidence in the function generator’s ability to functionally operate and  
meet specifications. These tests represent the absolute minimum set of  
performance checks recommended following any service activity.  
Auditing the function generator’s performance for the quick check points  
(designated by a Q) verifies performance for “normal” accuracy drift  
mechanisms. This test does not check for abnormal component failures.  
To perform the quick performance check, do the following:  
œ Set the function generator to the default power on state (power on default).  
A procedure is given on page 41.  
œ Perform a complete self-test. A procedure is given on page 21.  
4
œ Perform only the performance verification tests indicated with  
the letter Q.  
If the function generator fails the quick performance check, adjustment or  
repair is required.  
Performance Verification Tests  
The performance verification tests are recommended as acceptance tests  
when you first receive the function generator. The acceptance test results  
should be compared against the 1 year test limits. After acceptance, you  
should repeat the performance verification tests at every calibration interval.  
If the function generator fails performance verification, adjustment or  
repair is required.  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Frequency Verification  
Frequency Verification  
This test verifies the frequency accuracy of the two sources in the  
function generator. All output frequencies are derived from a single  
generated frequency, and only one frequency point is checked.  
The second test verifies the burst rate frequency.  
Set the function generator for each output indicated in the table below.  
Use a frequency meter to measure the output frequency. Compare the  
measured results to the test limits shown in the table. This is a 50W  
output termination test.  
Agilent 33120A  
Measurement  
OUT 1  
TERM  
BURST BURST  
Function  
Ampl  
Freq  
Nominal  
Error  
RATE  
CNT  
Q
Q
Sine wave  
3.5 Vrms 1.00 kHz  
1.00 kHz  
500 Hz  
50 W  
50 W  
0.02 Hz  
5 Hz  
Square wave  
3.5 Vrms 1.00 kHz 500 Hz 1 CYC  
Function Gain and Linearity Verification  
This test verifies the output amplitude accuracy specification for  
sine wave, triangle wave, ramp, and square wave outputs.  
Set the function generator for each output indicated in the table below.  
Use a DMM to measure the function generator ACrms output voltage.  
Compare the measured results to the test limits shown in the table.  
This is a HIGH Z output termination test.  
Agilent 33120A  
OUT TERM 1  
Ampl  
Measurement  
Function  
Sine wave  
Freq  
Nominal  
Error  
Q
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
7.0 Vrms  
5.7 Vrms  
5.7 Vrms  
5.7 Vrms  
10.0 Vrms  
8.0 Vrms  
1.0 kHz  
1.0 kHz  
100 Hz  
100 Hz  
100 Hz  
100 Hz  
7.0 Vrms  
5.7 Vrms  
5.7 Vrms  
5.7 Vrms  
10.0 Vrms  
8.0 Vrms  
0.07 Vrms  
0.057 Vrms  
0.057 Vrms  
0.057 Vrms  
0.1 Vrms  
0.08 Vrms  
Sine wave  
Triangle wave  
Ramp wave  
Square wave  
Square wave  
Q
1
Output termination set using front panel controls. HIGH Z assumes no load on  
output. 50W assumes a 50W – 0.1W load on output.  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
DC Function Offset Verification  
DC Function Offset Verification  
This test verifies the DC offset and DC output specifications.  
Set the function generator for each output indicated in the table below.  
Use a DMM to measure the function generator dcV output. Compare the  
measured results to the test limits shown in the table. This is a HIGH Z  
output termination test.  
Agilent 33120A  
OUT TERM 1  
HIGH Z  
Measurement  
Function  
DC Volts  
DC Volts  
Ampl  
Nominal  
Error  
Q
10.0 Vdc  
-10.0 Vdc  
10.0 Vdc  
-10.0 Vdc  
0.20 Vdc  
0.20 Vdc  
HIGH Z  
4
AC Amplitude Verification  
This procedure is used to check the output amplitude calibration of the  
function generator. Verification checks are performed to check the  
accuracy of the pre-attenuator and post attenuator. Make sure you have  
read “Test Considerations” on page 53.  
Set the function generator for each output indicated in the table on the  
next page. Use a DMM to measure the ACrms output voltage of the  
function generator. Compare the measured results to the test limits  
shown in the table. This is a HIGH Z output termination test.  
1
Output termination set using front panel controls. HIGH Z assumes no load on  
output. 50W assumes a 50W –0.1W load on output.  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
AC Amplitude Verification  
Agilent 33120A  
Measurement  
Function OUT TERM 1  
Ampl  
Freq  
Nominal  
Error  
Q
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
HIGH Z  
7.0 Vrms  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
1.00 kHz  
7.0 Vrms  
5.7 Vrms  
0.070 Vrms  
0.057 Vrms  
0.055 Vrms  
0.044 Vrms  
0.035 Vrms  
0.028 Vrms  
0.022 Vrms  
0.017 Vrms  
0.014 Vrms  
0.011 Vrms  
0.0088 Vrms  
0.0070 Vrms  
0.0055 Vrms  
0.0044 Vrms  
0.0035 Vrms  
0.0028 Vrms  
0.0022 Vrms  
0.0017 Vrms  
0.0014 Vrms  
0.0011 Vrms  
0.00088 Vrms  
0.00070 Vrms  
0.00055 Vrms  
0.00044 Vrms  
0.00036 Vrms  
5.7 Vrms  
5.5 Vrms  
5.5 Vrms  
4.4 Vrms  
4.4 Vrms  
3.5 Vrms  
3.5 Vrms  
2.8 Vrms  
2.8 Vrms  
2.2 Vrms  
2.2 Vrms  
1.7 Vrms  
1.7 Vrms  
1.4 Vrms  
1.4Vrms  
1.1 Vrms  
1.1 Vrms  
Q
0.88 Vrms  
0.70 Vrms  
0.55 Vrms  
0.44 Vrms  
0.35 Vrms  
0.28 Vrms  
0.22 Vrms  
0.17 Vrms  
0.14 Vrms  
0.11 Vrms  
0.088 Vrms  
0.070 Vrms  
0.055 Vrms  
0.044 Vrms  
0.036 Vrms  
0.88 Vrms  
0.70 Vrms  
0.55 Vrms  
0.44 Vrms  
0.35 Vrms  
0.28 Vrms  
0.22 Vrms  
0.17 Vrms  
0.14 Vrms  
0.11 Vrms  
0.088Vrms  
0.070 Vrms  
0.055 Vrms  
0.044 Vrms  
0.036 Vrms  
Q
Q
1
Output termination set using front panel controls. HIGH Z assumes no load on  
output. 50W assumes a 50W – 0.1W load on output.  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
AC Amplitude Verification  
Install the 50W feedthrough load between the DMM and the function  
generator output. Set the function generator for each output indicated in  
the table on the next page. Use a DMM to measure the ACrms output  
voltage of the function generator. Compare the measured results to the  
test limits shown in the table. This is a 50W output termination test.  
Agilent 33120A  
Ampl  
Measurement  
Function  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
OUT TERM 1  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
Freq  
Nominal  
Error  
Q
Q
3.5 Vrms  
2.8 Vrms  
2.2 Vrms  
1.7 Vrms  
1.4Vrms  
1.1 Vrms  
1.0000 kHz  
1.0000 kHz  
1.0000 kHz  
1.0000 kHz  
1.0000 kHz  
1.0000 kHz  
3.5 Vrms  
2.8 Vrms  
0.035 Vrms  
0.028 Vrms  
0.022 Vrms  
0.017 Vrms  
0.014 Vrms  
0.011 Vrms  
0.0088 Vrms  
0.0070 Vrms  
0.0055 Vrms  
0.0044 Vrms  
0.0035 Vrms  
0.0028 Vrms  
0.0022 Vrms  
0.0017 Vrms  
0.0014 Vrms  
0.0011 Vrms  
0.00088 Vrms  
0.00070 Vrms  
0.00055 Vrms  
0.00044 Vrms  
0.00035 Vrms  
0.00028 Vrms  
0.00022 Vrms  
0.00018 Vrms  
2.2 Vrms  
1.7 Vrms  
1.4Vrms  
1.1 Vrms  
0.88 Vrms 1.0000 kHz  
0.70 Vrms 1.0000 kHz  
0.55 Vrms 1.0000 kHz  
0.44 Vrms 1.0000 kHz  
0.35 Vrms 1.0000 kHz  
0.28 Vrms 1.0000 kHz  
0.22 Vrms 1.0000 kHz  
0.17 Vrms 1.0000 kHz  
0.14 Vrms 1.0000 kHz  
0.11 Vrms 1.0000 kHz  
0.088Vrms 1.0000 kHz  
0.070 Vrms 1.0000 kHz  
0.055 Vrms 1.0000 kHz  
0.044 Vrms 1.0000 kHz  
0.035 Vrms 1.0000 kHz  
0.028 Vrms 1.0000 kHz  
0.022 Vrms 1.0000 kHz  
0.018 Vrms 1.0000 kHz  
0.88 Vrms  
0.70 Vrms  
0.55 Vrms  
0.44 Vrms  
0.35 Vrms  
0.28 Vrms  
0.22 Vrms  
0.17 Vrms  
0.14 Vrms  
0.11 Vrms  
0.088Vrms  
0.070 Vrms  
0.055 Vrms  
0.044 Vrms  
0.035 Vrms  
0.028 Vrms  
0.022 Vrms  
0.018 Vrms  
4
Q
1
Output termination set using front panel controls. HIGH Z assumes no load on  
output. 50W assumes a 50W – 0.1W load on output.  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Amplitude Flatness Verification  
Amplitude Flatness Verification  
This test verifies the output amplitude flatness specification at selected  
frequencies. If you use a TVC (recommended) or a wide band ACrms  
voltmeter (with a 50W feed through load), perform this procedure as  
described. If you are using a measurement device that requires a transfer  
measurement (for example, a power meter), make the transfer in the  
reference measurement at 100 kHz.  
Set the function generator to the first output indicated in the table below  
and make a reference measurement. Select each function generator output  
in the table below and adjust the function generator output amplitude until  
the measured output is at the reference measurement. Compare the  
amplitude level set on the front panel to the test limits shown in the table.  
This test is a 50W output termination test.  
Agilent 33120A  
Measurement  
OUT 1  
Function  
Ampl  
Freq  
Nominal  
Error  
TERM  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
50 W  
Q
Q
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
3.0 Vrms  
3.0 Vrms  
3.0 Vrms  
3.0 Vrms  
3.0 Vrms  
3.0 Vrms  
3.0 Vrms  
3.0 Vrms  
3.0 Vrms  
3.0 Vrms  
3.0 Vrms  
1.00 kHz  
100.00 kHz  
500.00 kHz  
1.00 MHz  
3.00 MHz  
5.00 MHz  
7.00 MHz  
9.00 MHz  
11.00 MHz  
13.00 MHz  
15.00 MHz  
<reference>  
<reference>  
<reference>  
<reference>  
<reference>  
<reference>  
<reference>  
<reference>  
<reference>  
<reference>  
<reference>  
0.03 Vrms  
0.045 Vrms  
0.045 Vrms  
0.06 Vrms  
0.06 Vrms  
0.06 Vrms  
0.06 Vrms  
0.06 Vrms  
0.06 Vrms  
0.06 Vrms  
Q
1
Output termination set using front panel controls. HIGH Z assumes no load on  
output. 50W assumes a 50W –0.1W load on output.  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
AM Modulation Depth Verification  
AM Modulation Depth Verification  
This test verifies the modulation depth specification.  
Select each function generator output in the table below. Use a DMM to  
measure the function generator ACrms output voltage. Compare the  
measured results to the test limits shown in the table. This is a HIGH Z  
output termination test.  
Agilent 33120A  
Measurement  
AM Modulation  
Shape Freq  
OUT 1  
TERM  
Function  
Ampl  
Freq  
Depth Nominal  
Error  
Q
Sine wave HIGH Z 1.0 Vrms 1.00 kHz Sinewave 100 Hz  
Sine wave HIGH Z 1.0 Vrms 1.00 kHz Sinewave 100 Hz  
0%  
0.50 Vrms  
0.005 Vrms  
0.0061 Vrms  
100% 0.61 Vrms  
4
1
Output termination set using front panel controls. HIGH Z assumes no load on  
output. 50W assumes a 50W – 0.1W load on output.  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Optional Performance Verification Tests  
Optional Performance Verification Tests  
These tests are not intended to be performed with every calibration.  
They are provided as an aid for verifying additional instrument specifications.  
Square Wave Duty Cycle Verification  
This test verifies the duty cycle specification of the squarewave output.  
Select each function generator output in the table below. Use an integrating  
DMM to measure the Vdc output of the function generator. Compare the  
measured results to the test limits shown in the table. This is a HIGH Z  
output termination test.  
Agilent 33120A  
Measurement  
OUT 1  
TERM  
Duty  
Cycle  
Function  
Ampl  
Freq  
Nominal  
0.00 Vdc  
Error  
Square wave HIGH Z  
Square wave HIGH Z  
Square wave HIGH Z  
1.0 Vrms  
1.0 Vrms  
1.0 Vrms  
300.00 Hz  
300.00 Hz  
300.00 Hz  
50%  
0.020 Vdc  
0.020 Vdc  
0.020 Vdc  
25% - 0.50 Vdc  
75% + 0.50 Vdc  
The DMM used for this test must be an integrating multimeter. If the  
first step does not measure 0 Vdc, use an oscilloscope for this test.  
Do not use an auto-ranging function on the DMM for this test. Fix the  
DMM measurement range at 10 Vdc.  
1
Output termination set using front panel controls. HIGH Z assumes no load on  
output. 50W assumes a 50W – 0.1W load on output.  
62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Optional Performance Verification Tests  
Distortion Verification  
This test checks the Harmonic Distortion at selected frequencies and  
harmonics. This test requires the use of a spectrum analyzer with  
dynamic range, frequency range, and resolution bandwidth adequate for  
the measurement.  
Select each function generator output in the table below. Use a spectrum  
analyzer connected to the function generator output. Set the fundamental  
frequency reference to 0 dB and measure the 2nd through 5th harmonic  
frequencies relative to this reference. This test is a 50W output  
termination test.  
Agilent 33120A  
Measurement  
harmonic  
OUT 1  
TERM  
Value below  
reference  
Function  
Ampl  
Freq  
Fundamental 2nd  
3rd  
4th  
5th  
Sine wave  
Sine wave  
Sine wave  
Sine wave  
1.1 Vrms 20.00 kHz  
reference 40 kHz 60 kHz 80 kHz 100 kHz < 70 dB  
50 W  
50 W  
50 W  
50 W  
4
1.1 Vrms 100.00 kHz reference 200 kHz 300 kHz 400 kHz 500 kHz < 60 dB  
1.1 Vrms 1.00 MHz  
reference  
2 MHz 3 MHz 4 MHz 5 MHz  
< 45 dB  
< 35 dB  
1.1 Vrms 15.00 MHz reference 30 MHz 45 MHz 60 MHz 75 MHz  
1
Output termination set using front panel controls. HIGH Z assumes no load on  
output. 50W assumes a 50W – 0.1W load on output.  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Calibration Security Code  
Calibration Security Code  
This feature allows you to enter a security code (electronic key) to prevent  
accidental or unauthorized calibrations of the function generator.  
When you first receive your function generator, it is secured. Before you  
can adjust calibration constants you must unsecure the function generator  
by entering the correct security code. A procedure to unsecure the  
function generator is given on page 47.  
œ The security code is set to “HP033120” when the function generator is  
shipped from the factory. The security code is stored in non-volatile  
memory, and does not change when power has been off or after a remote  
interface reset.  
œ To secure the function generator from the remote interface, the security  
code may contain up to 12 alphanumeric characters as shown below.  
The first character must be a letter, but the remaining characters can be  
letters or numbers. You do not have to use all 12 characters but the first  
character must always be a letter.  
A _ _ _ _ _ _ _ _ _ _ _  
(12 characters)  
œ To secure the function generator from the remote interface but allow it to  
be unsecured from the front panel, use the eight-character format shown  
below. The first two characters must be “HP” and the remaining  
characters must be numbers. Only the last six characters are recognized  
from the front panel, but all eight characters are required. (To unsecure  
the function generator from the front panel, omit the “HP” and enter the  
remaining numbers.)  
H P _ _ _ _ _ _  
(8 characters)  
œ If you forget your security code, you can disable the security feature by  
adding a jumper inside the function generator, and then entering a  
new code. See the procedure on the following page.  
64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Calibration Security Code  
To Unsecure the Function Generator Without the Security Code  
To unsecure the function generator without the correct security code,  
follow the steps below. A procedure to unsecure the function generator is  
given on page 47. Also see “Electrostatic Discharge (ESD) Precautions” in  
chapter 6 before beginning this procedure.  
SHOCK HAZARD. Only service-trained personnel who are aware  
of the hazards involved should remove the instrument covers.  
The procedures in this section require that you connect the power  
cord to the instrument with the covers removed. To avoid  
electrical shock and personal injury, be careful not to touch the  
power-line connections.  
WARNING  
1 Disconnect the power cord and all input connections (front and rear terminals).  
2 Remove the instrument cover. Refer to the disassembly drawing on page 130.  
3 Connect the power cord and turn on the function generator.  
4
4 Apply a short between the two exposed metal pads on JM101 (located  
near U106 and U205) as shown in the figure below.  
5 While maintaining the short, enter any unsecure code. The function  
generator is now unsecured.  
6 Remove the short at JM101.  
7 Turn off and reassemble the function generator.  
8 The function generator is now unsecured and you can enter a new  
security code. Be sure you take note of the new security code.  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Calibration Count  
Calibration Count  
The calibration count feature provides an independent “serialization” of  
your calibrations. You can determine the number of times that your  
function generator has been calibrated. By monitoring the calibration  
count, you can determine whether an unauthorized calibration has been  
performed. Since the value increments by one for each calibration, a  
complete calibration increases the value by approximately 85 counts.  
œ The calibration count is stored in non-volatile memory and does not  
change when power has been off or after a remote interface reset.  
Your function generator was calibrated before it left the factory.  
When you receive your function generator, read the calibration count  
to determine its value.  
œ The calibration count increments up to a maximum of 32,767 after which  
it wraps around to 0. There is no way provided to program or reset the  
calibration count. It is an independent electronic calibration  
“serialization” value.  
Calibration Message  
You can use the calibration message feature to record calibration  
information about your function generator. For example, you can store  
such information as the last calibration date, the next calibration due  
date, the function generator’s serial number, or even the name and phone  
number of the person to contact for a new calibration.  
You can record information in the calibration message only from the  
remote interface. You can read the message from either the front-panel  
menu or the remote interface.  
œ The calibration message may contain up to 40 characters. The function  
generator can display up to 11 characters of the message on the front  
panel; any additional characters are truncated.  
œ The calibration message is stored in non-volatile memory, and does not  
change when power has been off or after a remote interface reset.  
66  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
General Calibration/Adjustment Procedure  
General Calibration/Adjustment Procedure  
The adjustment procedures described in chapter 4 use the CAL MENU to  
generate and set internal calibration constants. The general menu  
procedure is the same for all calibration setups. The following example  
demonstrates making the Frequency and Burst Rate adjustments.  
Shift  
1 Turn on the menu.  
Menu On/Off  
<
A: MOD MENU  
1
2 Move across to the CAL MENU choice on this level.  
F: CAL MENU  
4
3 Move down a level to the UNSECURED command.  
1: UNSECURED  
If the display shows SECURED, you will have to unsecure the function  
generator to perform the calibration procedures. A procedure is given on  
page 47.  
1
>
4
Move across to the CALIBRATE choice.  
2: CALIBRATE  
5 Move down one level.  
The display indicates the calibration setup number. You can change this  
number to perform individual specification adjustments.  
SETUP 00  
Enter  
6 Begin the Frequency and Burst Rate adjustment procedure.  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
General Calibration/Adjustment Procedure  
1
<
>
7 Move the flashing cursor over the digit to be edited.  
^
8 Change the value in the display to match the measured frequency.  
1.000,0040KHz  
9 Calculate and save the new value.  
CALIBRATING  
Enter  
10 Perform the next adjustment procedure.  
The setup number and function generator output is automatically set for  
the next adjustment procedure.  
SETUP ^01  
You will press ENTER twice for each calibration step, once to select  
the setup (as described in step 6) and once to enter the adjustment  
(as described in step 9).  
1
You can also use the knob to scroll left or right through the choices on each level of  
the menu  
68  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Aborting a Calibration in Progress  
Aborting a Calibration in Progress  
Sometimes it may be necessary to abort a calibration after the procedure  
has already been initiated. You can abort a calibration at any time by  
pressing any front-panel key (except Shift-Cancel ). When performing a  
calibration from the remote interface, you can abort a calibration by  
issuing a remote interface device clear message or by pressing the  
front-panel LOCAL key.  
Frequency and Burst Rate Adjustment  
The function generator stores two calibration constants related to  
frequency and burst rate output. The constants are calculated from the  
adjustment value entered and are stored at the completion of each setup.  
1 Use a frequency meter to measure the function generator output  
frequency for each setup in the following table. These adjustments use  
a 50W output termination.  
4
Nominal Output  
SETUP  
00 *  
FREQUENCY AMPLITUDE  
Adjustment for main frequency  
generator, sine wave output  
1.00 kHz  
500 Hz  
10 Vpp  
10 Vpp  
Adjustment for burst rate timing,  
pulse output.  
01  
* A new calibration (SETUP 86 – Rev 4.0) has been added as an alternative to SETUP 00.  
The new calibration outputs a 10 MHz sine wave, rather than the 1 kHz signal used for  
SETUP 00. The new calibration reduces slew rate dependent errors in the frequency  
measurement and is especially important when calibrating the Phase-Lock Assembly  
(Option 001). Note that either setup is sufficient to calibrate the carrier frequency and  
you don’t need to perform both.  
2 Use the CALIBRATE menu to adjust the displayed frequency at each  
setup to match the measured frequency and enter the value.  
3 Perform the Frequency Verification procedures on page 56.  
69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Function Gain and Linearity Adjustment  
Function Gain and Linearity Adjustment  
The function generator stores six calibration constants related to function  
gain and linearity. The constants are calculated from the adjustment  
value entered. If the calibration procedure is aborted before all setup steps  
have been completed, no calibration constants are stored.  
1 Use a DMM to measure the function generator ACrms output voltage for  
each setup in the following table. These adjustments use a HIGH Z  
output termination.  
Nominal Output  
SETUP  
02  
FREQUENCY  
1 kHz  
AMPLITUDE  
7.07 V rms  
5.6 V rms  
5.6 V rms  
5.6 V rms  
10.0 V rms  
1.1 Vrms  
Adjustment for sine wave gain.  
Adjustment for amplitude linearity.  
Adjustment for triangle wave gain.  
Adjustment for ramp gain.  
03  
1 kHz  
04  
100 Hz  
100 Hz  
100 Hz  
100 Hz  
05  
06  
Adjustment for square wave gain.  
Adjustment for square wave linearity.  
07  
2 Use the CALIBRATE menu to adjust the displayed amplitude at each  
setup to match the measured amplitude and enter the value.  
3 Perform the Function Gain and Linearity Verification procedures on page 56.  
AC Amplitude Adjustment (High-Z)  
The function generator stores twenty-two calibration constants related to  
HIGH Z output, and sixteen calibration constants related to 50W output.  
The constants are calculated from the adjustment value entered. The cal-  
ibration constants are stored following completion of setup 22 (HIGH Z  
output) and the calibration procedure may be aborted after that point.  
No calibration constants are stored if the procedures are aborted at any  
other setup.  
1 Use a DMM to measure the function generator ACrms output voltage for  
each setup in the following table. These adjustments use a HIGH Z  
output termination.  
70  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
AC Amplitude Adjustment (High-Z)  
Nominal Output  
SETUP  
8
FREQUENCY AMPLITUDE  
Adjustment for:  
2 dB Output Attenuator  
4 dB Output Attenuator  
6 dB Output Attenuator  
8 dB Output Attenuator  
10 dB Output Attenuator  
12 dB Output Attenuator  
14 dB Output Attenuator  
16 dB Output Attenuator  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
5.5 V rms  
4.4 V rms  
3.5 V rms  
2.8 V rms  
2.2 V rms  
1.7 V rms  
1.4 V rms  
1.1 V rms  
9
10  
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
21  
22  
23  
24  
25  
26  
27  
28  
29  
0.88 V rms 18 dB Output Attenuator  
0.70 V rms 20 dB Output Attenuator  
0.55 V rms 22 dB Output Attenuator  
0.44 V rms 24 dB Output Attenuator  
0.35 V rms 26 dB Output Attenuator  
0.28 V rms 28 dB Output Attenuator  
0.22 V rms 30 dB Output Attenuator  
4
5.5 V rms  
4.4 V rms  
3.5 V rms  
2.8 V rms  
2.2V rms  
1.7 V rms  
1.4 Vrms  
2 dB Pre-attenuator  
4 dB Pre-attenuator  
6 dB Pre-attenuator  
8 dB Pre-attenuator  
10 dB Pre-attenuator  
12 dB Pre-attenuator  
14 dB Pre-attenuator  
2 Use the CALIBRATE menu to adjust the displayed amplitude at each  
setup to match the measured amplitude and enter the value.  
3 Perform the AC Amplitude Verification procedures on page 57.  
71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Modulation Adjustment  
Modulation Adjustment  
The function generator stores three calibration constants related to  
amplitude modulation depth. The constants are calculated from the  
adjustment value entered. If the calibration procedure is aborted before  
all setup steps have been completed, no calibration constants are stored.  
1 Use a DMM to measure the function generator ACrms output voltage for  
each setup in the following table. These adjustments use a HIGH Z  
output termination.  
Nominal Output  
SETUP  
30  
FREQUENCY  
1 kHz  
AMPLITUDE  
3.5 Vrms  
Adjustment for:  
0% modulation depth.  
50% modulation depth.  
100% modulation depth.  
31  
1 kHz  
0.707 Vrms  
6.36 Vrms  
32  
1 kHz  
2 Use the CALIBRATE menu to adjust the displayed amplitude at each  
setup to match the measured amplitude and enter the value.  
3 Perform the AM Modulation Depth Verification procedures on page 61.  
NEW CALIBRATION: A new calibration (SETUP 85 Rev 4.0) has  
been added to eliminate a small residual error in the AM amplitude  
system which could potentially cause a failure of the AM amplitude  
verification. The new calibration operates just like the other AM  
calibrations (SETUP 30, 31, 32) in that the external measurement is  
AC Vrms with no load. The new calibration is not allowed until the  
other AM gain calibrations (SETUP 30, 31, 32) are performed.  
The new algorithm is designed such that the calibration should not be  
required again once the function generator has been calibrated at the  
factory. However, if you change any critical analog components which  
determine amplitude in AM modulation, you should perform the  
calibration again.  
72  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
AC Amplitude Adjustment (50W)  
AC Amplitude Adjustment (50W)  
1 The function generator stores 16 calibration constants related to 50W  
output. The constants are calculated from the adjustment value entered.  
The calibration constants are stored following completion of setup 49 and  
the calibration procedure may be aborted after that point. No calibration  
constants are stored if the procedures are aborted at any other setup.  
2
Use the DMM to measure the resistance of a 50W feedthrough load.  
Record the measurement for step 3. You can measure the load and cable  
resistance (recommended procedure) or just the load as shown below.  
4
3 Enter the following setup and use the calibrate menu to enter the  
measured value of the 50W feedthrough load (and cable). This number  
will be used to calculate the 50W output amplitude calibration constants.  
Nominal Input  
SETUP  
33  
LOAD Z  
Enter measured value of load.  
50 W  
Once the value of the 50W load and cable are entered, use the SAME  
load and cable for all 50W tests.  
73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
AC Amplitude Adjustment (50W)  
4 Use the DMM to measure the function generator ACrms output voltage  
for each setup in the table on the next page. These adjustments use the  
50W load and cable measured in step 2 and connected as shown below.  
74  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
AC Amplitude Adjustment (50W)  
Nominal Output  
FREQUENCY AMPLITUDE  
SETUP  
34  
Adjustment for:  
0 dB Output Attenuator  
2 dB Output Attenuator  
4 dB Output Attenuator  
6 dB Output Attenuator  
8 dB Output Attenuator  
10 dB Output Attenuator  
12 dB Output Attenuator  
14 dB Output Attenuator  
16 dB Output Attenuator  
18 dB Output Attenuator  
20 dB Output Attenuator  
22 dB Output Attenuator  
24 dB Output Attenuator  
26 dB Output Attenuator  
28 dB Output Attenuator  
30 dB Output Attenuator  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
1 kHz  
3.5 Vrms  
2.8 Vrms  
35  
36  
2.23 Vrms  
1.77 Vrms  
1.41 Vrms  
1.12 Vrms  
.887 Vrms  
.704 Vrms  
.559 Vrms  
.442 Vrms  
.350 Vrms  
.281 Vrms  
.223 Vrms  
.177 Vrms  
.141 Vrms  
.112 Vrms  
37  
38  
39  
40  
41  
42  
43  
44  
44  
4
46  
47  
48  
49  
5 Use the CALIBRATE menu to adjust the displayed amplitude at each  
setup to match the measured amplitude and enter the value.  
6 Perform the AC Amplitude Verification procedures beginning on page 57.  
75  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
DC Output Adjustment  
DC Output Adjustment  
The function generator stores nine calibration constants related to  
DC volts output. The constants are calculated from the adjustment value  
entered. The calibration constants are stored following completion of  
setup 59. No calibration constants are stored if the procedures are aborted  
at any other setup.  
1 Use a DMM to measure the function generator dcV output voltage for  
each setup in the following table. These adjustments use a HIGH Z  
output termination.  
Nominal Output  
SETUP  
50  
DC Volts  
- 8.0 Vdc  
8.0 Vdc  
0.0 Vdc  
0.0 Vdc  
0.0 Vdc  
0.0 Vdc  
0.0 Vdc  
0.0 Vdc  
0.0 Vdc  
0.0 Vdc  
Adjustment for:  
Negative offset gain  
51  
Positive offset gain  
52  
AM offset  
53  
2 dB Pre-attenuator offset.  
4 dB Pre-attenuator offset.  
6 dB Pre-attenuator offset.  
8 dB Pre-attenuator offset.  
10 dB Pre-attenuator offset.  
12 dB Pre-attenuator offset.  
14 dB Pre-attenuator offset.  
54  
55  
56  
57  
58  
59  
2 Use the CALIBRATE menu to adjust the displayed output voltage at  
each setup to match the measured voltage and enter the value.  
3 Perform the DC Function Offset Verification procedures on page 57.  
76  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Duty Cycle Adjustment  
Duty Cycle Adjustment  
The function generator stores two calibration constants related to  
squarewave offset and two calibration constants related to squarewave  
duty cycle. The constants are calculated from the adjustment value  
entered. The calibration constants are stored following completion of  
setup 63. No calibration constants are stored if the procedures are aborted  
at any other setup.  
1 Use a DMM to measure the function generator dcV output voltage for  
each setup in the following table. These adjustments use a HIGH Z  
output termination.  
For this test, the DMM must be set to a fixed range capable of measuring  
from +10 V to -10 V. Do not use an auto-ranging function for this test.  
4
Nominal Output  
SETUP  
60  
FREQUENCY AMPLITUDE  
10.0 Vdc  
-10.0 Vdc  
0.0 Vdc  
Positive squarewave offset.  
Negative squarewave offset.  
50% duty cycle squarewave.  
75% duty cycle squarewave  
61  
62  
300 Hz  
300 Hz  
63  
5.0 Vdc  
2 Use the CALIBRATE menu to adjust the displayed output voltage at  
each setup to match the measured voltage and enter the value.  
3 Perform the Squarewave Duty Cycle Verification procedures on page 62.  
AC Amplitude Flatness Adjustment  
The function generator stores eleven calibration constants related to AC  
Amplitude Flatness from 1 kHz to 15 MHz. The constants are calculated  
from the adjustment value entered and one of two calculation constants  
related to the type of measurement device you are using. The calibration  
constants are stored following completion of setup 82. No calibration  
constants are stored if the procedures are aborted at any other setup.  
77  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
AC Amplitude Flatness Adjustment  
This procedure can be performed with one of three types of measurement  
device; a broadband ACrms voltmeter, a power meter, or a thermal  
voltage converter. The procedure differs slightly depending upon the type  
of measurement device used. These adjustments us a 50W output termination.  
1 Use a DMM to measure the ACrms output voltage of the function generator  
and enter the measurement value for the setup in the table below.  
Nominal Output  
SETUP  
64  
FREQUENCY AMPLITUDE  
1 kHz 3.0 V rms  
Reference for:  
1 kHz flatness DAC gain  
2 a. If you are using a broadband ACrms voltmeter, proceed to step 3.  
b. If you are using a power meter capable of measurements at 1 kHz,  
use the power meter to measure the function generator output and enter  
the value for the setup in the table below. (If your power meter does not  
measure to 1 kHz, see the transfer measurement procedure below.)  
Nominal Output  
SETUP  
83  
FREQUENCY AMPLITUDE  
1 kHz 3.0 V rms  
Reference for:  
V rms, dBm  
Transfer Measurement Procedure  
If you are using a power meter not capable of measurement to 1 kHz,  
you can perform the transfer measurement at a different frequency.  
For example, the Agilent E4418A Power Meter with the Agilent 8482A  
probe and 20 dB attenuator are specified to a low frequency of 100 kHz.  
To use this measurement device, perform step 1, then use setup 65 to  
obtain a 100 kHz output. Measure the output with the power meter and  
record the measured value. Perform setup 83 and enter the recorded  
value (not a new measurement). Then, perform step 3 (you will use setup  
65 twice). This procedure assumes the output of the function generator is  
flat from 1 kHz to 100 kHz.  
c. If you are using a Thermal Voltage Converter (TVC), use the TVC to  
measure the function generator output and enter the measurement for  
the setup in the table below (TVC values entered are in mVdc).  
Nominal Output  
SETUP  
84  
FREQUENCY AMPLITUDE  
1 kHz 3.0 V rms  
Reference for:  
Thermal Voltage Converter  
78  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
AC Amplitude Flatness Adjustment  
3 For each setup in the table below, use the CALIBRATE command to  
change the displayed amplitude to match the measured amplitude.  
Nominal Output  
SETUP  
65  
FREQUENCY AMPLITUDE  
Adjustment for:  
100 kHz  
500 kHz  
1 MHz  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
3.0 V rms  
100 kHz amplitude flatness  
500 kHz amplitude flatness  
1 MHz amplitude flatness  
3 MHz amplitude flatness  
5 MHz amplitude flatness  
7 MHz amplitude flatness  
9 MHz amplitude flatness  
10 MHz amplitude flatness  
10.5 MHz amplitude flatness  
11 MHz amplitude flatness  
11.5 MHz amplitude flatness  
12 MHz amplitude flatness  
12.5 MHz amplitude flatness  
13 MHz amplitude flatness  
13.5 MHz amplitude flatness  
14 MHz amplitude flatness  
14.5 MHz amplitude flatness  
15 MHz amplitude flatness  
66  
67  
68  
3 MHz  
69  
5 MHz  
70  
7 MHz  
71  
9 MHz  
72  
10 MHz  
10.5 MHz  
11 MHz  
11.5 MHz  
12 MHz  
12.5 MHz  
13 MHz  
13.5 MHz  
14 MHz  
14.5 MHz  
15 MHz  
73  
74  
4
75  
76  
77  
78  
79  
80  
81  
82  
4 Perform the Amplitude Flatness Verification procedures on page 60.  
Shift  
Menu On/Off  
Completion of adjustment procedures. Return the function generator  
to the normal operating mode.  
EXITING  
79  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Output Amplifier Adjustment (Optional)  
Output Amplifier Adjustment (Optional)  
This adjustment procedure should only be performed following repairs to  
the Output Amplifier circuitry. The adjustment improves the high frequency  
performance of the Output Amplifier.  
1 Remove the function generator power and cover as described on page 130.  
2 Use a DMM to measure the ACrms voltage across J701 as shown below.  
Cable Shield is  
Circuit Ground  
3 Turn on the function generator.  
4 Set the function generator for a 1 kHz, 1V rms, sine wave output.  
5 Adjust R710 for a minimum reading on the voltmeter. Typical readings  
are less than 0.005 Vrms.  
6 Replace the covers as described on page 130.  
80  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Error Messages  
Error Messages  
The following tables are abbreviated lists of function generator’s error  
messages. They are intended to include errors which are likely to be  
encountered during the procedures described in this chapter. For a more  
complete list of error messages and descriptions, see chapter 5 in the  
Agilent 33120A User’s Guide.  
System Error Messages  
Error  
Error Message  
-330  
-350  
501  
502  
511  
512  
513  
514  
521  
522  
550  
Self-test Failed  
Too many errors  
Isolator UART framing error  
Isolator UART overrun error  
RS-232 framing error  
RS-232 overrun error  
4
RS-232 parity error  
Command allowed only with RS-232  
Input buffer overflow  
Output buffer overflow  
Command not allowed in Local  
Self-Test Error Messages  
Error  
601  
602  
603  
604  
605  
606  
607  
608  
625  
626  
627  
Error Message  
Front panel does not respond  
RAM read/write fail  
Waveform RAM readback failed  
Modulation RAM readback failed  
Serial configuration readback failed  
Waveform ASIC failed  
SYNC signal detection failure  
SYNC signal detection failure  
I/O Processor not responding  
I/O Processor failed self-test  
I/O Processor reset; possible low power line voltage  
81  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4 Calibration Procedures  
Error Messages  
Calibration Error Messages  
Error  
Error Message  
701  
702  
703  
704  
705  
706  
707  
708  
709  
760  
850  
851  
852  
853  
854  
855  
856  
857  
858  
Cal security disabled by jumper  
Cal secured  
Invalid secure code  
Secure code too long  
Cal aborted  
Cal value out of range  
Cal signal measurement out of range  
Flatness cal failed  
Cannot calibrate frequency while externally locked (Option 001)  
RAM checksum failure  
Cal setup invalid  
Negative offset gain cal required (CAL:SETup 50)  
Flatness DAC gain cal required (CAL:SETup 64)  
AM cal 1 required (CAL:SETup 30)  
AM cal 2 required (CAL:SETup 31)  
Cal load resistance not specified (CAL:SETup 33)  
Square wave positive offset cal required (CAL:SETup 60)  
Square wave 50% duty cycle cal required (CAL:SETup 62)  
AM cal 3 required (CAL:SETup 32)  
82  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5
5
Theory of  
Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Theory of Operation  
This chapter is organized to provide descriptions of the circuitry  
contained on the schematics shown in chapter 8. A block diagram  
overview is provided followed by more detailed descriptions of the  
circuitry contained in the schematics chapter.  
œ Block Diagram Overview . . . . . . . . . . . . . . . . 85  
œ Output Attenuator . . . . . . . . . . . . . . . . . . . 86  
œ Output Amplifier . . . . . . . . . . . . . . . . . . . . 87  
œ AM Modulation . . . . . . . . . . . . . . . . . . . . . 89  
œ Pre-attenuator . . . . . . . . . . . . . . . . . . . . . . 90  
œ Square Wave and Sync . . . . . . . . . . . . . . . . . 90  
œ Filters . . . . . . . . . . . . . . . . . . . . . . . . . . 92  
œ Waveform DAC/Amplitude Leveling/Waveform RAM . . 93  
œ Direct Digital Synthesis (DDS ASIC) . . . . . . . . . 95  
œ System DACs . . . . . . . . . . . . . . . . . . . . . . 96  
œ Floating Logic . . . . . . . . . . . . . . . . . . . . . . 97  
œ Earth-Referenced Logic . . . . . . . . . . . . . . . . . 98  
œ Power Supplies . . . . . . . . . . . . . . . . . . . . . 98  
œ Display and Keyboard . . . . . . . . . . . . . . . . . . 100  
The self-test procedures are described in chapter 6.  
84  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Block Diagram Overview  
Block Diagram Overview  
This discussion pertains to the block diagram shown on page 129.  
The function function generator’s circuitry is divided into two major  
blocks: the floating section and the earth (ground) reference section.  
All signal generation, control, and display functions are contained in  
the floating section. This section also contains the function generator’s  
main CPU.  
The floating section can be viewed in two pieces; the analog signal  
conditioning section (System DAC, Filters, Sync, Square wave,  
Pre-Attenuator, Output Amp, and Output Attenuator) and the digital  
logic section (Floating Logic, Digital Waveform Data Synthesis, and  
Waveform DAC).  
All signal generation, level control, and modulation functions are  
performed in the floating section. The waveform DAC generates two  
outputs, normal and inverted, between approximately 800 mVp-p and  
1 Vp-p. The DAC outputs are routed through anti-alias low-pass filters to  
eliminate higher frequency sampling products. The nominal x10 gain of  
the output amplifier, combined with preattenuator and output attenuator  
settings, are chosen such that the desired output amplitude is produced.  
The ground reference section uses a processor configured as a slave to the  
main CPU. This processor establishes external I/O communication with  
the main CPU through a bi-directional, optically isolated, serial  
communications link. The earth referenced processor controls low-level  
GPIB (IEEE-488) and RS-232 interface operation. The ground referenced,  
rear panel external trigger input uses a dedicated optical isolator to  
couple a trigger signal to the main CPU in the floating section.  
5
Separate power supplies are provided for the floating and ground  
reference sections. The front panel operates from the floating section with  
its logic common different from the CPU logic common.  
85  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Output Attenuator  
Output Attenuator  
Block 8 on block diagram page 129; Schematic on page 138.  
The Output Attenuator provides 0 to 30 dB of signal attenuation between  
the output amplifier section and the output BNC connector. Output signal  
levels are controlled by combining coarse amplitude control from the  
output attenuator section and pre-attenuator section with fine amplitude  
control from the Waveform DAC AMP_CTL signal.  
Four switched output attenuator pads are combined to achieve the desired  
signal attenuation as shown in the table below. Relays K801 through  
K804 either bypass an attenuator pad or select that attenuator. K801  
selects a 2 dB attenuator, K802 selects a 4 dB attenuator, K803 selects  
a 8 dB attenuator, and K804 selects a 16 dB attenuator. Relays are  
sequenced to provide signal attenuation in 6 dB steps. Intermediate  
amplitude levels are controlled by selecting 0 dB, 2 dB or 4 dB of signal  
attenuation through the pre-attenuator solid state switches in  
combination with reducing the output level of the waveform DAC itself.  
The AMP_CTL signal provides smoothly varying control of the Waveform  
DAC output level over a 0 dB to -2 dB range. This operation is described  
further in the Waveform DAC and System DAC discussions.  
Output Attenuation  
K801  
K802  
K803  
K804  
0 dB  
6 dB  
12 dB  
18 dB  
24 dB  
30 dB  
set  
reset  
set  
reset  
set  
set  
reset  
reset  
set  
set  
reset  
set  
set  
reset  
set  
reset  
reset  
set  
set  
set  
reset  
reset  
reset  
reset  
K801 through K804 are latching relays. Their set or reset state is selected  
by momentarily pulsing the appropriate coil of the relay. Relay coils are  
pulsed with 5 volts for 15 ms through relay drivers U301 and U302.  
The main controller, U102, writes data bytes to ASIC U103 which  
transmits this data to the relay drivers via the internal 3-wire serial data bus  
(SERCLK, SERDAT, and SERSTB) to accomplish the relay state changes.  
A 30 MHz filter, composed of L801, C801, and C802, eliminates wideband  
noise from the function generator output. The output amplifier and output  
attenuators are protected from damage by clamps CR801 and CR802 and  
by fuse F801. The function generator is protected from accidental  
application of voltages <10 volts for short durations.  
86  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Output Amplifier  
Output Amplifier  
Block 7 on block diagram page 129; Schematic on page 137.  
The output amplifier drives the function generator’s signal output  
through the output attenuator section. The output amplifier exhibits an  
approximate 35 MHz bandwidth and 1000 V/ms slew rate. AC signals  
originating from the DAC+ and DAC- signal paths are combined at the  
input of the amplifier. The output amplifier exhibits a nominal x(-10)  
voltage gain from its -AMP_IN input and a nominal x12 voltage gain  
from its +AMP_IN input. A dc offset signal, related to the front panel  
output offset value, is also summed with the ac signal at the input of the  
amplifier. A simplified block diagram of the output amplifier is shown below.  
5
87  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Output Amplifier  
The block diagram shows four basic stages: dc amplifier, input differential  
amplifier, gain, and power output. The amplifier’s input differential  
amplifier stage and gain stage are symmetrical. The +AMP_IN and  
-AMP_IN inputs are both amplified through complementary amplifiers  
whose outputs are summed together at the input of the power output  
stage. Transistors Q701, Q702, Q704 and Q707 form the complementary  
input differential amplifiers. Q708 and Q705 are current sources which  
provide bias to the input differential amplifiers. Q709 and Q710 are  
emitter follower amplifiers used to couple the respective differential  
amplifier outputs to the gain stage transistors Q711 and Q715 which  
provide virtually all of the amplifiers open loop gain (~ x1000).  
The power output stage is a wideband, class C buffer amplifier. Emitter  
followers Q714 and Q716 buffer the gain stage output from loading by the  
power output emitter follower transistors Q713 and Q718. Idle current  
bias for these power output transistors is set by the ratios of R732, R726  
and transistor matching between Q713, Q714 and their equivalents in  
the other half of the stage: R734, R727 and Q718, Q716. Transistors Q712  
and Q717 are current sources which provide bias to emitter followers  
Q714 and Q716 respectively.  
The low frequency and dc performance of the amplifier is controlled by  
U702. This amplifier is used to sense the dc offset present at the  
+AMP_IN and -AMP_IN inputs and servo the output amplifier dc offset  
to zero volts; to the limit of U702’s own dc offset performance. U702 also  
provides a means to add a desired dc offset value into the output signal  
path through the x (-1) gain of the OUT_OFFSET signal.  
The output amplifier employs a current feedback technique to set the  
closed-loop gain. The emitters of Q701 and Q702 are the virtual summing  
node points in the amplifier. Amplifier closed loop gain is controlled  
predominately by the following ratios:  
2
R740  
R715  
R710  
R716  
R717  
2
R740  
R719  
R710  
R720  
R711  
and  
Variable resistor R710 is used to match the gain through the high frequency  
feedback path (described above) and the dc feedback path summed  
through resistors R705, R706. The feedback signal current is injected into  
the amplifier through the emitters of Q701 and Q702 respectively.  
88  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
AM Modulation  
AM Modulation  
Blocks 3 and 6 on block diagram page 129; Schematics on pages 136 and 133.  
Amplitude modulation is performed by analog multiplier U603 combining  
the AM_IN and +FUNCTION and -FUNCTION signals. Modulation depths  
from 0% to 120% are set by varying the signal at AM_IN.  
When the amplitude modulation function is selected, the output of U603  
is switched into the +AMP_IN signal path by K602. At the same time,  
the -AMP_IN signal path is grounded, cutting the output signal  
amplitude in half, to accommodate the more than two times peak signal  
levels required by >100% modulation depth.  
5
The AM_IN signal is a combination of any external modulation inputs  
applied to the rear panel BNC connector and the internally generated  
AM signals. The function generator can internally synthesize an 8-bit  
modulation wave shape through DAC U313. Data from any standard or  
arbitrary wave shape can be used as the modulating wave shape.  
Modulating wave shapes are automatically expanded or compressed in  
length, as required, to meet the specified modulating frequency setting.  
Changes in the function generator output will lag changes in the  
modulating frequency because new modulation data must be computed  
and downloaded internally for every frequency change.  
The AM_GAIN and AM_OFFSET dc signals are used to calibrate and vary  
the am modulation depth settings. AM_GAIN controls the peak-to-peak  
output level from U313 in response to modulation depth setting changes.  
Likewise, the AM_OFFSET signal varies inversely to the AM_GAIN signal,  
as the AM depth setting is varied, to produce a constant signal offset in the  
composite AM_IN modulation signal. The net AM_IN offset is independent  
of the modulating ac signal component or AM depth setting.  
89  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Pre-attenuator  
Pre-attenuator  
Block 6 on block diagram page 129; Schematic on page 136.  
All signals, except square waves, pass through the preattenuator.  
The preattenuator multiplexes eight resistive 2 dB attenuators to provide  
attenuation from 0 dB to 14 dB in 2 dB steps. The 0 dB, 2 dB, and 4 dB  
attenuation steps are used for level settings between the 6 dB steps  
selected in the output attenuator section. Amplitude settings between  
these 2 dB steps are set by smoothly varying the Waveform DAC output  
level from 0 dB to -2 dB of its nominal level via the AMP_CTL signal.  
Output attenuator 6 dB steps, preattenuator 0 dB, 2 dB, and 4 dB steps,  
and small variations (0 dB to 2 dB) of the Waveform DAC output level  
are combined to produce each amplitude setting.  
In the preattenuator, U601 and U602 are operated as 8-to-1 multiplexers,  
each providing selectable 2 dB attenuation steps. Because of the gain  
imbalance of the output amplifier (x 12 on +AMP_IN and x (-10) on  
-AMP_IN), the +signal path U601 has an additional 2 dB attenuation  
always present (R601 and R602) to equalize the nominal gains in both  
the plus and minus signal paths.  
Square Wave and Sync  
Block 6 on block diagram page 129, schematic on page 136.  
During square wave outputs, a sine wave signal is generated internally  
and squared-up by comparator U620. Square wave amplitude control is  
accomplished by variable gain amplifier Q603 and Q604 and switched  
into the output signal path through relay K601. A simplified diagram of  
the square wave generator is shown below.  
90  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Square Wave and Sync  
Transistors Q601 and Q602 buffer the output of the sine wave anti-alias  
filter to the input of comparator U620. Square wave duty cycles are  
controlled by the SQ_SYM input on the inverting input of the comparator.  
The squarewave outputs of U620 are amplified by variable gain  
amplifiers Q603 and Q604.  
The amplifier gain output level is controlled by the variable current  
source Q605 and U307D in response to the System DAC dc signal  
SW_AMP. Squarewave variable gain amplifier output signal levels are  
unbalanced by resistors R643 and R644 to correct for the output  
amplifier + and - gain differences as discussed in the preattenuator  
section on page 90.  
Latching relay K601 connects the square wave into the +FUNCTION  
and -FUNCTION paths. The relay set or reset state is selected by  
momentarily pulsing the appropriate coil. Relay coils are pulsed with  
5 volts for 15 ms through relay driver U301. The main controller, U102,  
writes data bytes to ASIC U103 which transmits this data to the relay  
drivers via the internal 3-wire serial data bus (SERCLK, SERDAT,  
and SERSTB) to accomplish relay state changes.  
Multiplexer U604 selects one of five sources for the SYNC output: off,  
modulation sync, square wave comparator output, RUN*, or Arbitrary  
waveform sync. ARB_SYNC is derived from the WA14 line through  
U210B. U215B and U210C control the pulse width of the ARB_SYNC  
(arbitrary waveform sync) signal. Square wave sync is taken from the  
inverting output of square wave comparator U620. U620 also generates  
the MOD_SYNC (modulation sync) through U217. Buffer U621 inverts  
the sync signal and provides the output current drive to the SYNC output  
BNC connector.  
5
91  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Filters  
Filters  
Block 5 on block diagram page 129; Schematic on page 135.  
The output of the Waveform DAC passes through one of two anti-alias  
filters. A 17 MHz 9th order elliptical filter is used for the sine wave and  
square wave output functions. A 10 MHz 7th order Bessel filter is used  
for filtering all other output functions, including all arbitrary waveshapes.  
The diagrams below show the typical frequency response of these filters.  
The filters are switched in or out of the signal path by latching relays  
K501 and K502. Their set or reset state is selected by momentarily  
pulsing the appropriate coil of the relay. Relay coils are pulsed with  
5 volts for 15 ms through relay drivers U301 and U302. The main  
controller, U102, writes data bytes to ASIC U103 which transmits this data  
to the relay drivers via the internal 3-wire serial data bus (SERCLK,  
SERDAT, and SERSTB) to accomplish these relay state changes. When  
K501 and K502 are set, the 10 MHz Bessel filter is selected.  
92  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Waveform DAC/Amplitude Leveling/Waveform RAM  
Waveform DAC/Amplitude Leveling/Waveform RAM  
Block 4 on block diagram page 129; Schematic on page 134.  
The Waveform DAC, U407, converts 12-bit digital data from waveform  
RAM’s U404 and U405 into positive and negative analog voltages.  
A simplified diagram of the Waveform DAC circuitry is shown below.  
5
The preattenuator, filters, and associated circuits in the output signal path  
provide an approximate 25W load for the Waveform DAC. The Waveform  
DAC nominally produces a 40 mA differential output current — yielding  
differential 1 Vac output signals. Wave shape (amplitude) data is loaded  
into the waveform RAM by the main controller CPU U102. Once loaded,  
these data are addressed by the DDS ASIC. The rate at which addresses  
are incremented determines the output waveform frequency. Waveform  
RAM output data is latched and shifted to ECL levels by U402 and U403  
for input to the waveform digital-to-analog converter (DAC) U407. DDS  
ASIC U206, waveform data latches U402 and U403, and the Waveform  
DAC U407 are clocked at 40 MHz. The 40 MHz clock is generated by  
oscillator U413 and ECL level-shifter U401.  
93  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Waveform DAC/Amplitude Leveling/Waveform RAM  
The Waveform DAC voltage reference is driven by U410B. This reference  
controls the magnitude of the nominal 0 to -40 mA DAC output current.  
The reference level is varied to produce 0 to -2 dB fine amplitude level  
control via dc signal AMP_CTL and 2 dB of dynamic amplitude flatness  
correction for static and swept frequency operation via flatness correction  
dac U409. These reference voltage adjustments are summed together in  
amplifier U410B. Amplitude flatness correction data are stored in  
calibration memory during calibration. These data are used to produce a  
modulation program with corresponding 8-bit amplitude correction data  
values which are gated to latch U412 during operation. These data  
provide real-time correction of the output amplitude level as frequency  
changes are made. Amplifier U408 and Q401 use the waveform DAC  
reference voltage to center the waveform DAC output signal near 0 volts.  
U404 and U405 are the high-speed waveform RAM. Together, U404 and  
U405 form a 16383 x 12-bit RAM. Each RAM stores and outputs 6 bits of  
the waveform DAC 12-bit WD data bus. RAM U404 drives the least  
significant 4 bits and U405 drive the most significant 8 bits of the WD  
data bus. Note that DAC U407 calls D1 the most-significant bit (MSB)  
and D12 it’s least-significant bit (LSB). Waveform RAM addresses are  
controlled by the DDS ASIC’s WA (waveform address) bus.  
94  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Direct Digital Synthesis (DDS ASIC)  
Direct Digital Synthesis (DDS ASIC)  
Block 2 on block diagram page 129; Schematic on page 132.  
The DDS ASIC, U206, controls the WA (waveform address) and MA  
(modulation address) busses. The waveform address is used by the  
waveform RAMs U404 and U405. The modulation data bus is used by the  
modulation RAM U205.  
The DDS ASIC is comprised of several internal registers and addressing  
state machines. Instructions are written to the DDS ASIC by the main CPU  
via memory mapped control registers U108 and U202. When loading data  
into Waveform RAM or Modulation RAM, addresses on the WA and MA  
busses are incremented by ASIC U206. ASIC addresses are incremented  
by each rising edge of the TRIG line while writing data into these RAM.  
The state of the HOST_RQ* line controls whether the main CPU or the  
modulation RAM U205 is sourcing instructions to the DDS ASIC internal  
state machines. The Modulation RAM is loaded with frequency values  
and amplitude flatness correction values or AM modulation data for latch  
U309 and AM dac U313. Data multiplexer U217 and flip-flop U215 are  
used to preselect and synchronize the modulation sync source available to  
the SYNC output terminal multiplexer U604.  
The external trigger input OGEXT is optically isolated by U213 and  
applied to an input of trigger source multiplexer U214. The external  
trigger input is used for triggering the start of a frequency sweep or  
burst output and for externally gating the output signal on and off  
asynchronously. U214 selects one of seven trigger sources for use by  
U206 for initiating its internal program.  
5
95  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
System DACs  
System DACs  
Block 3 on block diagram page 129; Schematic on page 133.  
All output amplitudes are derived from the internal voltage reference of  
System DAC U303. The system dac track/hold amplifier outputs are used  
to provide controllable bias voltages to various analog circuits including  
AM modulation depth, square wave amplitude, square wave duty cycle,  
output dc offset, and output amplitude level. The System DAC is  
programmed and responds to the main controller via the internal 3-wire  
serial data bus SERCLK, SERRBK, and SERSTB. The System DAC is  
multiplexed to 7 track/hold amplifiers through U304. Each track/ hold  
amplifier is refreshed approximately every 3 ms to maintain its output  
setting. Changes to track/hold amplifier outputs are accomplished by  
dwelling on that position for an extended period.  
96  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Floating Logic  
Floating Logic  
Block 1 on block diagram page 129; Schematic on page 131.  
The floating logic controls the operation of the entire function function  
generator. All output functions and bus command interpretation is  
performed by the main CPU, U102. The front panel and earth referenced  
logic operate as slaves to U102. The main CPU portion of the floating logic  
section is clocked from a 12 MHz ceramic resonator, Y101. Non-volatile  
EEPROM U106 stores arbitrary waveform data, calibration constants,  
calibration secure code, calibration count, and last instrument state.  
The main CPU, U102, is a 16-bit micro controller. The 16-bit A (address)  
bus and 8-bit AD (address/data) bus are used to provide digital  
communication with the 256k byte program ROM U104, 32k byte RAM  
U105, 128k byte non-volatile EEPROM U106, 32k byte high speed  
Modulation RAM U205, 16k x 12-bit high speed waveform RAM U404  
and U405 and DDS ASIC U206.  
Gate array U103 provides CPU address latching and memory mapping  
functions. There are four internal registers in U103: a configuration  
register, an 8-bit counter register, a serial transmit/receive register, and  
an internal status register. RAM chip select signal RAMCE* and CPU  
port bits RAMA13 and RAMA14 are used to access 4 - 8k byte banks of  
program data RAM. Similarly, 4 banks of 56k non-volatile EEPROM and  
2 banks of 56k non-volatile RAM are gated from CPU port bits PRG16,  
PRG17, and WAVA16 and U103 signal ROMCE*. Addresses on the CPU  
address bus are valid when the ALE line is high. Memory mapping of  
control of registers U107 and U202, DDS ASIC U206, data transceivers  
U201, U203, U204, and write enables for RAM U404 and U405 and U205  
are controlled by data selector U108.  
5
The U103 serial register controls the front panel, relay drivers U301 and  
U302, and System DAC U303 through a serial data bus. Front panel  
signals are FPDI, FPSK*, and FPDO. Interrupts from the front panel are  
detected by U103 and signaled to U102 by CHINT. The FPINT line from  
U102 signals the front panel that U103 has data to send. The internal  
3-bit serial data bus (U102) uses SERCK, SERDAT, and SERSTB to send  
data to various registers. SERRBK (serial read back) is used by self test  
to verify operation of U103, relay drivers U301, U302, and System DAC  
shift register U305.  
97  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Earth-Referenced Logic  
The main CPU, U102, communicates with the earth referenced logic  
through an optically isolated asynchronous serial data link. U101 isolates  
the incoming data (OG_RXD*) from the earth referenced logic. Similarly,  
U901 isolates the data from U102 (OG_TXD) to the earth reference logic.  
Data is sent in an 11-bit frame at a rate of 187.5 k bits/second. When the  
RS-232 interface is selected, data is sent across the serial link at 93.75 k  
bits/second. The 11-bit internal data frame is configured for one start bit,  
eight data bits, one control bit, and one stop bit.  
Earth-Referenced Logic  
Block 9 on block diagram page 129; Schematic on page 139.  
The earth referenced section provides all rear panel input/output  
capability. Microprocessor U903 handles GPIB (IEEE-488) control  
through bus interface chip U904 and bus receiver/driver chips U907 and  
U908. The RS-232 interface is also controlled through U903. RS-232  
transceiver chip U906 provides the required level shifting to approximate  
9 volt logic levels through on-chip charge-pump power supplies using  
capacitors C904 and C906. Communication between the earth referenced  
logic interface circuits and the floating logic is accomplished through an  
optically-isolated bi-directional serial interface. Isolator U101 couples  
data from U903 to microprocessor U102. Isolator U901 couples data from  
U102 to microprocessor U903.  
Power Supplies  
Block 10 on block diagram page 129; Schematic on page 140.  
The power supply section, is divided into two isolated blocks similar to  
the floating logic and earth referenced logic sections discussed earlier.  
The floating supply outputs are 18 Vdc, +5 Vdc, -5.2 Vdc (VEE), and a  
6 Vrms center tapped filament supply for the vacuum fluorescent display.  
All earth referenced logic is powered from a single +5 Vdc supply.  
Power-on reset signals are provided by both the floating and earth  
referenced power supplies. In addition, the floating section +5 Vdc supply  
incorporates a power failure detection circuit which provides a priority  
interrupt signal to the main CPU (U102).  
98  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Power Supplies  
The ac mains are connected by a fused power entry module, P1.  
This module incorporates the functions of mains connection, on/off  
switching, fusing, and line voltage selection (100/120/220 (230)/240).  
The line voltage selection function of module P1 selects which primary  
winding of power transformer T1 is energized. The transformer  
secondary windings are connected to the main pc board through  
connector J1001.  
The floating +5 Vdc and -5.2 Vdc supplies are produced by a bridge  
rectifier formed by diodes CR1006 through CR1009, filter capacitors  
C1005 and C1009, and voltage regulators U1003 and U1004. U1005 and  
CR1011 form a clamp circuit to provide over voltage protection in the  
event of a mains or transformer failure. The PFAIL and PONRST*  
signals are derived from the floating +5 Vdc supply. PFAIL is asserted  
when the raw 5 Vdc supply drops below 6.4 V signaling an unstable  
power supply condition to the main CPU (U102). Current instrument  
state information is copied to non-volatile RAM, U106 for future recall if  
needed. The PONRST* signal holds the main CPU and other logic in a  
reset state until after the +5 Vdc logic power supply is fully operational.  
This signal is generally active only following application of line power to  
the instrument.  
The floating 18 volt supplies are produced by bridge rectifier CR1001,  
filter capacitors C1001 and C1003, and regulators U1001 and U1002.  
These supplies are used to power all analog circuits in the function generator.  
In addition, the vacuum fluorescent display is driven from the 18 volt  
supplies. A separate winding of T1 provides a center tapped 6 Vrms  
filament supply for the display. Bias circuit CR1010, R1009, and C1011  
generate the required cathode dc bias for the display filament supply.  
5
The 5 volt earth referenced supply is produced by rectifier CR1051,  
C1053, and regulator U1051. This supply is earth referenced through  
the screw used to mount the PC board to the instrument chassis.  
The GPIB (IEEE-488) and RS-232 computer interfaces and the rear-  
panel EXT Trigger circuits are powered from this supply. A controlled  
power-on reset signal for processor U903 is generated by U1052.  
99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 5 Theory of Operation  
Display and Keyboard  
Display and Keyboard  
Block 11 on block diagram page 129; Schematic on page 141.  
The front panel circuits consist of vacuum fluorescent display control,  
display high voltage drivers, and keyboard scanning. Communication  
between the front panel and floating logic circuits is accomplished  
through a 4-wire bi-directional serial interface. The main CPU, U102,  
can cause a hardware reset to processor U1101 by signal IGFPRES.  
The front panel logic operates from -13 volts (logic 1) and -18 volts  
(logic 0). The four serial communication signals are level shifted by  
comparator U1301 from the floating logic 0 V to 5 V levels to the -18 V  
to -13 V levels present on the front panel assembly. The front panel logic  
high supply (-13 volts) is produced from the -18 volt supply by voltage  
regulator U1102.  
Display anode and grid voltages are +18 volts for an on segment and  
-18 volts for an off segment. The -12 V cathode bias for the display is  
provided by the main pc board’s filament winding center tap bias circuit  
CR1010, R1009, and C1011 shown on the power supply schematic  
(see page 140).  
Keyboard scanning is accomplished through a conventional scanned  
row-column key matrix. Keys are scanned by outputting data at  
microprocessor U1101 port pins P0.0 through P0.4 to poll each key  
column for a key press. Column read-back data are read by the  
microprocessor at port pins P1.0 through P1.3 for decoding and  
communication to the floating logic circuits. Rotary knob quadrature  
inputs are read directly by the microprocessor port pins P1.6 and P1.7.  
100  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6
6
Service  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service  
This chapter discusses the procedures involved for returning a failed  
function generator to Agilent for service or repair. Subjects covered  
include the following:  
œ Operating Checklist . . . . . . . . . . . . . . . . . . . 103  
œ Types of Service Available . . . . . . . . . . . . . . . 104  
œ Repackaging for Shipment . . . . . . . . . . . . . . . 105  
œ Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . 105  
œ Electrostatic Discharge (ESD) Precautions . . . . . . 106  
œ Surface Mount Repair . . . . . . . . . . . . . . . . . . 106  
œ To Replace the Power-Line Fuse . . . . . . . . . . . . 107  
œ To Replace the Output Protection Fuse (F801) . . . . 107  
œ Troubleshooting Hints . . . . . . . . . . . . . . . . . 108  
œ Self-Test Procedures . . . . . . . . . . . . . . . . . . 110  
102  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6 Service  
Operating Checklist  
Operating Checklist  
Before returning your function generator to Agilent for service or repair,  
check the following items:  
Is the function generator inoperative?  
Verify that the ac power cord is connected to the function generator.  
Verify that the front-panel Power switch is depressed.  
Verify that the power-line fuse is good (see page 22).  
The function generator is shipped from the factory with a  
500 mAT fuse installed. This is the correct fuse for all line voltages.  
Verify the power-line voltage setting.  
See “To prepare the function generator for use” on page 21.  
Does the function generator fail self-test?  
Verify that the correct power-line voltage is selected.  
See “To prepare the function generator for use” on page 21.  
Is the function generator’s output inoperative?  
Turn off the function generator and remove the power cord.  
Using an ohmmeter, measure the resistance between the output  
BNC center conductor and case. If the ohmmeter measures >100W,  
the internal output protection fuse, F801, may be open.  
6
103  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6 Service  
Types of Service Available  
Types of Service Available  
If your function generator fails during the warranty period (within three  
years of original purchase), Agilent will replace or repair it free of charge.  
After your warranty expires, Agilent will repair or replace it at a  
competitive price. The standard repair process is “whole unit exchange”.  
The replacement units are fully refurbished and are shipped with new  
calibration certificates.  
Standard Repair Service (worldwide)  
Contact your nearest Agilent Technologies Service Center. They will  
arrange to have your function generator repaired or replaced.  
Agilent Express Unit Exchange (U.S.A. only)  
You will receive a refurbished, calibrated replacement Agilent 33120A in  
1 to 4 days.  
1
2
Call 1-877-447-7278 and ask for “Agilent Express”.  
œ You will be asked for your serial number, shipping address, and a credit  
card number to guarantee the return of your failed unit.  
œ If you do not return your failed unit within 15 business days, your credit  
card will be billed for the cost of a new 33120A.  
Agilent will immediately send a replacement 33120A directly to you.  
œ The replacement unit will come with instructions for returning your  
failed unit. Please retain the shipping carton and packing materials to  
return the failed unit to Agilent. If you have questions regarding these  
instructions, please call 1-877-447-7278.  
œ The replacement unit will have a different serial number than your  
failed unit. If you need to track your original serial number, a blank  
label will be shipped with the replacement unit to record your original  
serial number.  
104  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6 Service  
Repackaging for Shipment  
Repackaging for Shipment  
If the unit is to be shipped to Agilent for service or repair, be sure to:  
œ Attach a tag to the unit identifying the owner and indicating the required  
service or repair. Include the instrument model number and your full  
serial number.  
œ Place the unit in its original container with appropriate packaging  
material.  
œ Secure the container with strong tape or metal bands.  
If the original shipping container is not available, place your unit in a  
container which will ensure at least 4 inches of compressible packaging  
material around all sides for the instrument. Use static-free packaging  
materials to avoid additional damage to your unit.  
Agilent suggests that you always insure shipments.  
Cleaning  
Clean the outside of the instrument with a soft, lint-free, slightly  
dampened cloth. Do not use detergent. Disassembly is not required  
or recommended for cleaning.  
6
105  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6 Service  
Electrostatic Discharge (ESD) Precautions  
Electrostatic Discharge (ESD) Precautions  
Almost all electrical components can be damaged by electrostatic  
discharge (ESD) during handling. Component damage can occur at  
electrostatic discharge voltages as low as 50 volts.  
The following guidelines will help prevent ESD damage when servicing  
the function generator or any electronic device.  
œ Disassemble instruments only in a static-free work area.  
œ Use a conductive work area to dissipate static charge.  
œ Use a conductive wrist strap to dissipate static charge accumulation.  
œ Minimize handling.  
œ Keep replacement parts in original static-free packaging.  
œ Remove all plastic, styrofoam, vinyl, paper, and other static-generating  
materials from the immediate work area.  
œ Use only anti-static solder suckers.  
SHOCK HAZARD. Only service-trained personnel who are aware  
of the hazards involved should remove the instrument covers.  
To avoid electrical shock and personal injury, make sure to  
disconnect the power cord from the function generator before  
removing the covers.  
WARNING  
Surface Mount Repair  
Surface mount components should only be removed using soldering irons  
or desoldering stations expressly designed for surface mount components.  
Use of conventional solder removal equipment will almost always result in  
permanent damage to the printed circuit board and will void your Agilent  
factory warranty.  
106  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6 Service  
To Replace the Power-Line Fuse  
To Replace the Power-Line Fuse  
The power-line fuse is located within the function generator’s fuse-holder  
assembly on the rear panel (see page 22). The function generator is shipped  
from the factory with a 500 mAT slow-blow fuse installed (part number  
2110-0458). This is the correct fuse for all line voltages.  
To Replace the Output Protection Fuse (F801)  
The Output Protection Fuse is located inside the function generator.  
This fuse is a thru-hole soldered 500 mA part (part number 2110-0716).  
The fuse is located near the output connector (J801) on the main PC board.  
You will need to disassemble the function generator to replace this fuse  
(use a TORX T-15 driver to remove the screws located on the rear panel).  
The disassembly procedure is shown on page 130.  
6
107  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6 Service  
Troubleshooting Hints  
Troubleshooting Hints  
This section provides a brief checklist of common failures. Before trouble-  
shooting or repairing the function generator, make sure the failure is in  
the instrument rather than any external connections. Also make sure  
that the instrument is accurately calibrated. The function generator’s  
circuits allow troubleshooting and repair with basic equipment such as a  
1
6 -digit multimeter and a 100 MHz oscilloscope.  
2
Unit is Inoperative  
Verify that the ac power cord is connected to the function generator.  
Verify that the front-panel Power switch is depressed.  
Verify that the power-line fuse is good (see page 22).  
The function generator is shipped from the factory with a  
500 mAT fuse installed. This is the correct fuse for all line voltages.  
Verify the power-line voltage setting.  
See “To prepare the function generator for use” on page 21.  
Unit Reports Error 760  
This error may be produced if you accidentally turn off power the unit  
during a calibration or while changing a non-volatile state of the  
instrument. Recalibration or resetting the state should clear the error.  
If the error persists, a hardware failure may have occurred.  
Unit Fails Self-Test  
Verify that the correct power-line voltage setting is selected. Also, ensure  
that all terminal connections (both front panel and rear terminals) are  
removed while the self-test is performed. Failure of the System DAC U03  
will cause many self-test failures.  
108  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6 Service  
Troubleshooting Hints  
Power Supply Problems  
SHOCK HAZARD. Only service-trained personnel who are aware  
of the hazards involved should remove the instrument covers.  
The procedures in this section require that you connect the power  
cord to the instrument with the covers removed. To avoid  
electrical shock and personal injury, be careful not to touch the  
power-line connections.  
WARNING  
Check that the input to the supply voltage regulator is at least  
1 volt greater than its output.  
Circuit failures can cause heavy supply loads which may pull  
down the regulator output voltage.  
Check the main supply voltages as tabulated below.  
Power Supply  
+5 Ground Ref.  
+5 Floating  
Minimum  
4.75V  
Maximum  
5.25V  
4.75V  
5.25V  
-5.2 Floating  
+18 Floating  
-18 Floating  
-5.46V  
17.0V  
-4.94V  
19.1V  
-19.1V  
4.75V  
-17.0V  
5.25V  
+5REF Floating  
Some circuits produce their own local power supplies derived from  
the main supplies. Be sure to check that these local supplies are  
active. In particular, the output amplifier and front panel sections  
have local supplies. Always check that the power supplies are free  
of ac oscillations using an oscilloscope.  
6
109  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6 Service  
Self-Test Procedures  
Self-Test Procedures  
Power-On Self-Test  
Each time the function generator is powered on, a small set of self-tests  
are performed. These tests check that the minimum set of logic and  
measurement hardware are functioning properly. The power-on self-test  
performs checks 601, 625, and 626.  
Complete Self-Test  
Hold the shift key while turning on the power to perform a complete  
self-test. The tests are performed in the order shown below.  
Performing Individual Tests  
You can perform individual self-tests through the SYStem menu and  
TEST command. The parameters allowed are ALL, 603, 604, 605, 606,  
607, and 608. All numbered tests are looped to give a continuous pass/fail  
indication when started from the menu (they will repeat the waveform or  
sync signal until interrupted).  
601  
Front panel does not respond The main CPU U102 attempts to  
establish serial communications with the front panel processor U1101.  
During this test, U1101 turns on all display segments. Communication  
must function in both directions for this test to pass. If this error is  
detected during power-up self-test, the function generator will beep twice.  
This error is only readable from the remote interface.  
602  
603  
RAM read/write failed This test writes and reads a 55 and AA  
H
H
checker board pattern to each address of ram U103. Any incorrect  
readback will cause a test failure.  
Waveform RAM readback failed This test writes and reads a A55  
H
and 5AA checker board pattern (12-bit) to each address of waveform  
H
ram U404 and U405. The test writes and reads the pattern twice,  
at alternating addresses. Any incorrect readback will cause a test failure.  
604  
Modulation RAM readback failed This test writes and reads a 5A  
H
and A5 checker board pattern (8-bit) to each address of Modulation ram  
H
U205. The test writes and reads the pattern twice, at alternating  
addresses. Any incorrect readback will cause a test failure.  
110  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 6 Service  
Self-Test Procedures  
605  
606  
Serial configuration readback failed This test re-sends the last 3  
byte serial configuration data to all the serial path (SERDAT, SERBCK,  
SERCLK). The data is then clocked back into U103 and compared against  
the original 3 bytes sent. A failure occurs if the data do not match.  
This tests checks the serial data path through U301, U302, and U305.  
This tests does not check the serial path to the system DAC U303.  
Waveform ASIC failed This test is the first part of test 607 (below).  
This test sets up a burst modulation program of a special waveform  
(a four period ramp wave) where the data at each waveform address is  
the same as the last 12 bits of its address. The burst modulation  
waveform is run for 1 cycle plus 1 address and the waveform data is read  
back and compared. A correct result infers that the modulation program  
ran correctly and halted at the correct address. This checks the ability of  
U205 and U206 to correctly run a burst modulation program. This test  
and test 604 give a high confidence in the modulation circuitry. This test  
will fail if a trigger signal is present on the rear panel Ext Trig BNC input.  
607  
608  
625  
SYNC signal detection failure, Bessel filter path This test runs the  
special waveform described in test 606 and counts transitions of the  
SYNC line. The test should provide 8 transitions of the SYNC signal.  
The test also checks the 7th Order Bessel filter path and U620 and U604.  
An incorrect number of transitions will generate an error. This test will  
fail if a trigger signal is present on the rear panel Ext Trig BNC input.  
SYNC signal detection failure, Elliptical filter path This test runs  
a special waveform and counts transitions of the SYNC line. The test  
should provide 2 transitions of the SYNC line. The test checks the 9th  
Order Elliptical filter path. An incorrect number of transitions will  
generate an error. This test will fail if a trigger signal is present on the  
rear panel Ext Trig BNC input.  
6
I/O processor does not respond This test checks that communications  
can be established between U102 and U903 through the optically isolated  
(U101 and U901) serial data link. Failure to establish communication in  
either direction will generate an error. If this condition is detected at  
power-on self-test, the function generator will beep and the error  
annunciator will be on.  
626  
701  
I/O processor failed self-test This test causes the earth referenced  
processor U903 to execute an internal, ram test. Failure will generate  
an error.  
Checks that the calibration security disable jumper is removed. If the jumper  
is shorted at power on, all non-volatile RAM is reset to initial factory values.  
111  
Download from Www.Somanuals.com. All Manuals Search And Download.  
112  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7
7
Replaceable Parts  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Replaceable Parts  
This chapter contains information to help you order replacement parts for  
your 33120A Function Generator. The parts lists are divided into the  
following groups:  
œ 33120-66521 Main PC Assembly (A1)  
œ 33120-66502 Front-Panel Display and Keyboard PC Assembly (A2)  
œ 33120A Mainframe  
œ Manufacturer’s List  
Parts are listed in alphanumeric order according to their schematic  
reference designators. The parts lists include a brief description of the  
part with applicable Agilent part number and manufacturer part number.  
To Order Replaceable Parts  
You can order replaceable parts from Agilent using the Agilent part  
number or directly from the manufacturer using the manufacturer’s part  
number. Note that not all parts listed in this chapter are available as  
field-replaceable parts. To order replaceable parts from Agilent, do the  
following:  
1 Contact your nearest Agilent Sales Office or Agilent Service Center.  
2 Identify the parts by the Agilent part number shown in the replaceable  
parts list. Note that not all parts are directly available from Agilent;  
you may have to order certain parts from the specified manufacturer.  
3 Provide the instrument model number and serial number.  
114  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
33120-66521 – Main PC Assembly  
33120-66521 – Main PC Assembly  
Reference  
Designator  
Agilent Part  
Number  
Mfr.  
Code  
Mfr. Part  
Number  
Qty  
Part Description  
C101-C102  
C103-C105  
C107  
0160-6497  
0160-5945  
0160-5945  
0160-6497  
0160-5945  
0160-6497  
0160-5945  
0160-6497  
0160-5945  
0160-6497  
78  
26  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V  
04222 12065C104KAT A  
04222 08055C103KAT A  
04222 08055C103KAT A  
04222 12065C104KAT A  
04222 08055C103KAT A  
04222 12065C104KAT A  
04222 08055C103KAT A  
04222 12065C104KAT A  
04222 08055C103KAT A  
04222 12065C104KAT A  
C108  
C109-C110  
C111-C113  
C114-C117  
C201-C202  
C203  
C204  
C205  
0160-5945  
0160-6497  
0160-5945  
0160-6497  
0160-6497  
0160-6497  
0160-5945  
0160-6497  
0160-5945  
0160-6497  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V  
04222 08055C103KAT A  
04222 12065C104KAT A  
04222 08055C103KAT A  
04222 12065C104KAT A  
04222 12065C104KAT A  
04222 12065C104KAT A  
04222 08055C103KAT A  
04222 12065C104KAT A  
04222 08055C103KAT A  
04222 12065C104KAT A  
C206-C207  
C208-C210  
C211-C213  
C301-C306  
C308-C309  
C310-C312  
C313-C318  
C319  
C320  
C401-C402  
C404  
C405  
C406  
C407  
C408  
C409  
C410  
C411  
0160-6497  
0160-5967  
0160-5962  
0160-5967  
0160-5962  
0160-5967  
0160-6497  
0160-5945  
0160-6497  
0160-5945  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 100 pF 50 V  
CAP-FXD 15 pF 50 V  
CAP-FXD 100 pF 50 V  
CAP-FXD 15 pF 50 V  
CAP-FXD 100 pF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.01 uF 50 V  
04222 12065C104KAT A  
04222 08051A101JAT A  
04222 08051A150JAT A  
04222 08051A101JAT A  
04222 08051A150JAT A  
04222 08051A101JAT A  
04222 12065C104KAT A  
04222 08055C103KAT A  
04222 12065C104KAT A  
04222 08055C103KAT A  
6
2
C412  
C413-C422  
C423  
C424  
C425  
C426-C427  
C428  
C429-C432  
C503-C504  
C505-C506  
C507-C508  
0160-6497  
0160-5959  
0160-5947  
0160-6497  
0160-6736  
0160-5967  
0160-6497  
0160-7405  
0160-5953  
0160-5964  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 33 pF 50 V  
CAP-FXD 1000 pF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.01uF +-10% 50 V CER X7R  
CAP-FXD 100 pF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 560 pF +-10% 50 V CER X7R  
CAP-FXD 270 pF 50 V  
CAP-FXD 180 pF 50 V  
04222 12065C104KAT A  
04222 08051A330JAT A  
04222 08055C102KAT A  
04222 12065C104KAT A  
51406 GRM426X7R103K50  
04222 08051A101JAT A  
04222 12065C104KAT A  
28480 0160-7405  
1
4
2
2
5
3
02444 08051A271JAT A  
04222 08051A181JATRA  
7
C509-C510  
C521-C522  
C531-C532  
C533-C534  
0160-5958  
0160-5954  
0160-5955  
0160-5976  
2
3
4
2
CAP-FXD 39 pF 50 V  
CAP 220 pF 5% 50V  
CAP-FXD 68 pF 50 V  
CAP-FXD 12 pF 50 V  
04222 08051A390JAT A  
04222 08051A221JAT A  
04222 08051A680JATRA  
04222 08051A120JAT A  
115  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
33120-66521 – Main PC Assembly  
Reference  
Designator  
Agilent Part  
Number  
Mfr.  
Code  
Mfr. Part  
Number  
Qty  
Part Description  
C535-C536  
C539-C540  
C541-C542  
C543-C544  
C545-C546  
C547-C548  
C551-C552  
C553-C554  
C601-C607  
C608  
0160-5952  
0160-7721  
0160-5965  
0160-5967  
0160-7733  
0160-5953  
0160-5961  
0160-5965  
0160-6497  
0160-5954  
2
2
2
CAP-FXD 330 pF 50 V  
04222 08055A331JAT A  
04222 08055A820FATMA  
04222 08051A151JAT A  
04222 08051A101JAT A  
04222 08055A101FAT_A  
04222 08051A271JAT A  
04222 08051A220JAT A  
04222 08051A151JAT A  
04222 12065C104KAT A  
04222 08051A221JAT A  
CAP-FXD 82 pF +-1% 50 V CER C0G  
CAP-FXD 150pF +-5% 50 V CER C0G  
CAP-FXD 100 pF 5%  
CAP-FXD 100 pF +-1% 50 V CER C0G  
CAP-FXD 270 pF 50 V  
CAP-FXD 22 pF 50 V  
CHIP CAPACITOR  
CAP-FXD 0.1 uF 25 V  
CAP 220 pF 5% 50V  
2
2
C623  
C624  
C630  
C631  
C632-C633  
C702  
C703-C704  
C705  
C706-C707  
C709  
0180-3975  
0160-5945  
0160-5945  
0160-5955  
0160-5957  
0160-5953  
0160-5975  
0160-5942  
0180-3975  
0160-5975  
5
2
CAP-FXD 2.2 uF 20 V TA  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 68 pF 50 V  
CAP-FXD 47 pF  
CAP-FXD 270 pF 50 V  
CAP-FXD 10 pF 50 V  
CAP-FXD 1 pF 50 V  
CAP-FXD 2.2 uF 20 V TA  
CAP-FXD 10 pF 50 V  
04222 TAJB225M020  
04222 08055C103KAT A  
04222 08055C103KAT A  
04222 08051A680JATRA  
04222 08051A470JAT A  
02444 08051A271JAT A  
04222 08051A100JAT A  
04222 08051A1R0CAT A  
04222 TAJB225M020  
3
2
04222 08051A100JAT A  
C710  
C711  
C712  
C713  
C714-C719  
C720  
C802  
C901  
C902  
0180-3975  
0180-3859  
0180-3975  
0180-3859  
0160-6497  
0160-5942  
0160-5964  
0160-6497  
0160-5955  
0160-6497  
CAP-FXD 2.2 uF 20 V TA  
CAP-FXD 1000 uF+-20% 35 VDC  
CAP-FXD 2.2 uF 20 V TA  
CAP-FXD 1000 uF+-20% 35 VDC  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 1 pF 50 V  
CAP-FXD 180 pF 50 V  
CAP-FXD 0.1 uF 25 V  
04222 TAJB225M020  
S4217 SME35VB102M12.5X25LL  
04222 TAJB225M020  
S4217 SME35VB102M12.5X25LL  
04222 12065C104KAT A  
04222 08051A1R0CAT A  
04222 08051A181JATRA  
04222 12065C104KAT A  
04222 08051A680JATRA  
04222 12065C104KAT A  
2
CAP-FXD 68 pF 50 V  
CAP-FXD 0.1 uF 25 V  
C903-C906  
C908  
C909-C911  
C913  
C1001  
C1002  
C1003  
C1004  
C1005  
C1006  
0160-5945  
0160-6497  
0160-6497  
0180-4313  
0180-3751  
0180-4313  
0180-3751  
0180-4086  
0180-3751  
0160-6497  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 0.1 uF 25 V  
CAP 2200 uF 50V  
CAP-FXD 1 uF 35 V TA  
CAP-FXD 2200 uF 50V  
CAP-FXD 1 uF 35 V TA  
CAP-FXD .01 F+-20% 25 VDC AL  
CAP-FXD 1 uF 35 V TA  
CAP-FXD 0.1 uF 25 V  
04222 08055C103KAT A  
04222 12065C104KAT A  
04222 12065C104KAT A  
S4217 KME50VB222M18X35LL  
04222 TAJB105M035  
S4217 KME50VB222M18X35LL  
04222 TAJB105M035  
S4217 SME25VN103M22X45LL  
04222 TAJB105M035  
04222 12065C104KAT A  
2
3
2
C1007-C1008  
C1009  
C1010-C1011  
C1013  
C1014  
0180-4589  
0180-4116  
0160-6497  
0180-4116  
0160-6497  
1
4
CAP-FXD +-20% 16 V AL-ELCTLT  
CAP-FXD 22 uF 20 V TA  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 22 uF 20 V TA  
CAP-FXD 0.1 uF 25 V  
S4217 SMH16VN153M22X35LL  
04222 TAJD226M020  
04222 12065C104KAT A  
04222 TAJD226M020  
C1015-1016  
04222 12065C104KAT A  
116  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
33120-66521 – Main PC Assembly  
Reference  
Designator  
Agilent Part  
Number  
Mfr.  
Code  
Mfr. Part  
Number  
Qty  
Part Description  
C1017-1019  
C1020-1021  
C1051-C1052  
C1053  
C1054  
C1055  
0160-5947  
0160-5945  
0160-6497  
0180-4086  
0180-4116  
0160-6497  
CAP-FXD 1000 pF 50 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD .01 F+-20% 25 VDC  
CAP-FXD 22 uF 20 V TA  
CAP-FXD 0.1 uF 25 V  
04222 08055C102KAT A  
04222 08055C103KAT A  
04222 12065C104KAT A  
S4217 SME25VN103M22X45LL  
04222 TAJD226M020  
04222 12065C104KAT A  
CR301  
CR302-CR303  
CR401  
1906-0291  
1902-1541  
1902-1541  
1906-0291  
1902-1512  
1902-1610  
1902-1610  
1901-1386  
1902-1807  
1906-0291  
8
3
DIODE- 70V 100MA  
04713 MBAV99  
DIODE-ZNR 3.3V 5% TO-236 (SOT-23)  
DIODE-ZNR 3.3V 5% TO-236 (SOT-23)  
DIODE- 70V 100MA  
DIODE-ZNR 7.5V 5% PD=.155W TC=+5.3MV  
DIODE-ZNR SMB 3.3V 5% 1W  
DIODE-ZNR SMB 3.3V 5% 1W  
DIODE SI PN SOT23 100V 750MA  
DIODE-ZNR SMB 24V 42MA 1.0W 5%  
DIODE- 70V 100MA  
04713 BZX84C3V3  
04713 BZX84C3V3  
04713 MBAV99  
04713 BZX84C7V5  
04713 1SMB5913B  
04713 1SMB5913B  
28480 1901-1386  
04713 1SMB5934BT3  
04713 MBAV99  
CR601  
CR701-CR702  
CR703-CR704  
CR709-CR710  
CR801-CR802  
CR803-CR804  
CR901-CR904  
2
4
2
2
8
CR1001  
1906-0407  
1906-0291  
1902-1609  
1901-1607  
1902-1609  
1906-0407  
1901-1346  
2
DIODE-FW BRDG 400V 1A  
DIODE- 70V 100MA  
DIODE-ZNR 6.2V 5% PD=1.5W IR=5UA  
DIODE-PWR RECT SMC 400V 2.5A 2.5US S3G  
DIODE-ZNR 6.2V 5% PD=1.5W IR=5UA  
DIODE-FW BRDG 400V 1A  
71744 DF04S  
CR1002-1003  
CR1004-1005  
CR1006-1009  
CR1010  
CR1051  
CR1052  
04713 MBAV99  
04713 1SMB5920B  
71744 S3G  
04713 1SMB5920B  
71744 DF04S  
3
4
1
DIODE,TVS,D0214AB,43V,1500WP,SMCJ43CA  
71744 SMCJ43CA  
E901  
F801  
9164-0173  
2110-0716  
9170-1421  
9170-1421  
9170-1421  
9170-1421  
1205-0880  
1205-0213  
1205-0213  
1
1
7
BEEPER  
51406 PKM24-4AO-1  
75915 R251.500T1  
28480 9170-1421  
28480 9170-1421  
28480 9170-1421  
28480 9170-1421  
13103 7021B-TC10-MT  
13103 2228B  
FUSE-SUBMINIATURE .50A 125V NTD AX UL  
BEAD, SHIELDING (CHOKE)  
BEAD, SHIELDING (CHOKE)  
BEAD, SHIELDING (CHOKE)  
BEAD, SHIELDING (CHOKE)  
HEATSINK-TO-220  
FB401-FB402  
FB601  
FB701-FB702  
FB1001-1002  
HS1001-1004  
HSQ713  
4
2
HEAT SINK SGL TO-5/TO-39-CS  
HEAT SINK SGL TO-5/TO-39-CS  
HSQ718  
13103 2228B  
J101  
J102  
J301  
J401  
J601  
J701  
J801  
J901  
J902  
J903  
J1001  
J1002  
J1051  
1252-4484  
1252-4683  
1250-1884  
1250-0257  
1251-2969  
1251-5066  
1251-2969  
1252-2266  
1252-2161  
1250-1884  
1252-5136  
1251-5066  
1252-4487  
1
1
2
1
2
2
CONN-POST TYPE 2.0-PIN-SPCG 12-CONT  
CONN-FRCC VERT MALE 10PIN SMC  
CONNECTOR-RF BNC RCPT, 50-OHM  
CONNECTOR-RF SMB PLUG 50-OHM  
CONN PHONE VERT  
27264 52007-1210  
76381 N3662-6202  
00779 227161-6  
00779 413990-3  
27264 15-24-0503  
27264 22-04-1021  
27264 15-24-0503  
00779 748959-1  
00779 554923-2  
00779 227161-6  
27264 26-64-4090  
27264 22-04-1021  
27264 26-64-4030  
CONN DIS VERT MALE 2PIN FP  
CONN PHONE VERT  
1
1
CONN-RECT D-SUBMIN 9-CKT 9-CONT  
CONN-RECT MICRORBN 24-CKT 24-CONT  
CONN-RF BNC RCPT PC-W-STDFS 50-OHM  
CONN-DIS FRIC LOCK VERT MALE 9PIN  
CONN DIS VERT MALE 2 PIN FP  
7
1
1
CONN-POST TYPE .156-PIN-SPCG 3-CONT  
117  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
33120-66521 – Main PC Assembly  
Reference  
Designator  
Agilent Part  
Number  
Mfr.  
Code  
Mfr. Part  
Number  
Qty  
Part Description  
JM1001-1004  
JM1051  
0699-1503  
0699-1503  
7
RESISTOR .05 +-100% TKF  
RESISTOR .05 +-100% TKF  
28480 0699-1503  
28480 0699-1503  
K501-K502  
K601-K602  
K801-K804  
0490-1664  
0490-1664  
0490-1664  
8
RELAY 2C 5VDC-COIL 1A 110VDC  
RELAY 2C 5VDC-COIL 1A 110VDC  
RELAY 2C 5VDC-COIL 1A 110VDC  
28480 0490-1664  
28480 0490-1664  
28480 0490-1664  
L501-L502  
L503-L504  
L505-L506  
L521-L522  
L531-L532  
L533-L534  
L535-L536  
L537-L538  
L601-L602  
L801  
9140-1716  
9140-1102  
9140-1425  
9140-1102  
9140-1103  
9140-1102  
9140-1101  
9140-1102  
0699-1503  
9140-1099  
2
8
2
INDUCTOR 910 nH +2% -2%  
INDUCTOR 560 nH +-5% 2.8W-MMX3.4LG-MM 24226 03273  
INDUCTOR SMT 270 nH 5% 24226 SM3-270J  
INDUCTOR 560 nH +-5% 2.8W-MMX3.4LG-MM 24226 03273  
INDUCTOR 680 nH +-5% 2.8W-MMX3.4LG-MM 24226 03273  
INDUCTOR 560 nH +-5% 2.8W-MMX3.4LG-MM 24226 03273  
INDUCTOR 470 nH +-5% 2.8W-MMX3.4LG-MM 24226 03273  
INDUCTOR 560 nH +-5% 2.8W-MMX3.4LG-MM 24226 03273  
RESISTOR .05 +-100% TKF  
INDUCTOR SMT 220 nH 5%  
02113 1008CS-911XGBC  
2
2
28480 0600-1503  
09021 KL32TER22J  
1
4
Q401  
1854-1037  
1854-1037  
1853-0516  
1853-0567  
1854-1148  
1853-0516  
1854-1148  
1853-0567  
1853-0516  
1854-1037  
TRANSISTOR PD=350 MW FT=300 MHZ  
TRANSISTOR PD=350 MW FT=300 MHZ  
TRANSISTOR PD=350 MW FT=600 MHZ  
TRANSISTOR PD=350 MW FT=250 MHZ  
TRANSISTOR PD=350 MW FT=650 MHZ  
TRANSISTOR PD=350 MW FT=600 MHZ  
TRANSISTOR PD=350 MW FT=650 MHZ  
TRANSISTOR PD=350 MW FT=250 MHZ  
TRANSISTOR PD=350 MW FT=600 MHZ  
TRANSISTOR PD=350 MW FT=300 MHZ  
04713 MMBT3904  
04713 MMBT3904  
04713 MMBTH81  
04713 MMBT3906  
04713 MMBTH10  
04713 MMBTH81  
04713 MMBTH10  
04713 MMBT3906  
04713 MMBTH81  
04713 MMBT3904  
Q601-Q602  
Q603-Q604  
Q605  
Q701  
Q702  
Q704  
Q705  
Q707  
Q708  
4
2
2
Q709  
Q710  
Q711-Q712  
Q713  
Q714  
Q715-Q717  
Q718  
1854-1445  
5063-1420  
1853-0728  
1854-0597  
1853-0728  
1854-1303  
1853-0293  
1
1
3
1
TRANSISTOR NPN SI SC-59  
TRANSISTOR PNP 600 MHZ  
28480 1854-1445  
28480 5063-1420  
28480 1853-0728  
04713 2N5943  
28480 1853-0728  
28480 1854-1303  
04713 2N5583  
TRANSISTOR PNP SI TO-261AA (SOT-223)  
TRANSISTOR NPN 2N5943 SI TO-39 PD=1W  
TRANSISTOR PNP SI TO-261AA (SOT-223)  
TRANSISTOR NPN SI TO-261AA (SOT-223)  
TRANSISTOR P RF TO39 30V 500 MA 1GHZ  
3
R101-R104  
R106-R109  
R110  
R111-R112  
R114-R116  
R117  
R118-R119  
R120-R121  
R201  
0699-1318  
0699-1423  
0699-1386  
0699-1318  
0699-1391  
0699-1423  
0699-1318  
0699-1423  
0699-1318  
0699-1391  
0699-3034  
20  
12  
1
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
RESISTOR 215 +-1% .125 W TKF TC=0+-100  
RESISTOR 5.62K +-1% .125 W TKF TC=0+-100 28480 0699-1386  
28480 0699-1318  
28480 0699-1423  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
RESISTOR 10K +-1% .125 W TKF TC=0+-100  
RESISTOR 215 +-1% .125 W TKF TC=0+-100  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
RESISTOR 215 +-1% .125 W TKF TC=0+-100  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
RESISTOR 10K +-1% .125 W TKF TC=0+-100  
RESISTOR 1K +-1% .1W TKF TC=0+-100  
28480 0699-1318  
28480 0699-1391  
28480 0699-1423  
28480 0699-1318  
28480 0699-1423  
28480 0699-1318  
28480 0699-1391  
2M627 MCR10-FZHM-F-1001  
11  
R203  
R205  
1
118  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
33120-66521 – Main PC Assembly  
Reference  
Designator  
Agilent Part  
Number  
Mfr.  
Code  
Mfr. Part  
Number  
Qty  
Part Description  
R301  
R302  
R303-R304  
R305  
R307-R308  
R309  
R310  
R311  
R313-R314  
0699-1403  
0699-3211  
0699-1318  
0699-1391  
0699-3431  
0699-1937  
0699-1384  
0699-1344  
0699-1434  
1
2
RESISTOR 31.6K +-1% .125 W TKF TC=0+-100 28480 0699-1403  
RESISTOR 39.2K 1% 1206 .125 W 200V TC=100  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
RESISTOR 10K +-1% .125 W TKF TC=0+-100  
RESISTOR 4.99K +-1% .125 W TKF TC=0+-100  
28480 0699-3211  
28480 0699-1318  
28480 0699-1391  
28480 0699-3431  
6
1
6
4
2
RESISTOR 1.24K 1% 1206 .125 W TC=100 200V 28480 0699-1937  
RESISTOR 4.64K +-1% .125 W TKF TC=0+-100  
RESISTOR 10 +-1% .125 W TKF TC=0+-100  
RESISTOR 619 +-1% .125 W TKF TC=0+-100  
28480 0699-1384  
28480 0699-1344  
28480 0699-1434  
R316  
R317  
R318-R320  
R321  
R322  
R323  
R324  
R401  
R402  
0699-1391  
0699-1374  
0699-3431  
0699-1318  
0699-3211  
0699-1400  
0699-1432  
0699-1344  
0699-3698  
0699-3041  
RESISTOR 10K +-1% .125 W TKF TC=0+-100  
RESISTOR 1.78K +-1% .125 W TKF TC=0+-100  
RESISTOR 4.99K +-1% .125 W TKF TC=0+-100  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
RESISTOR 39.2K 1% 1206 .125 W 200V TC=100  
RESISTOR 26.1K +-1% .125 W TKF TC=0+-100  
RESISTOR 511 +-1% .125 W TKF TC=0+-100  
RESISTOR 10 +-1% .125 W TKF TC=0+-100  
RESISTOR 10 +-1% .1W TKF TC=0+-100  
RESISTOR 3.48K +-1% .1W TKF TC=0+-100  
28480 0699-1391  
28480 0699-1374  
28480 0699-3431  
28480 0699-1318  
28480 0699-3211  
28480 0699-1400  
28480 0699-1432  
28480 0699-1344  
2M627 MCR10-F-10R0  
2M627 MCR10-F-3481  
2
1
8
1
1
R404  
R405-R406  
R407-R408  
R409  
R410  
R411  
R412  
R413  
R414  
R415  
0699-2889  
0699-2832  
0699-1381  
0699-2489  
0699-1415  
0699-2937  
0699-2431  
0699-1318  
0699-1391  
0699-3594  
2
2
1
3
19  
1
1
RESISTOR 4.7 +-5% .125 W TKF  
RESISTOR 200 +-0.1% .125 W TF TC=0+-25  
RESISTOR 3.48K +-1% .125W TKF TC=0+-100 2M627 MCR18-F-3481  
RESISTOR 10K +-0.1% .125 W TF TC=0+-25  
RESISTOR 100 +-1% .125 W TKF TC=0+-100  
RESISTOR 16K +-0.1% .125W TF TC=0+-25  
RESISTOR 3.32K +-1% .125 W TKF TC=0+-100  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
RESISTOR 10K +-1% .125 W TKF TC=0+-100  
RESISTOR 20.5K +-0.1% .125W TF TC=0+-25  
2M627 MCR18J4R  
11502 W1206R032000BT  
11502 W1206R031002BT  
28480 0699-1415  
09021 RN73E2BTE1602B  
28480 0699-2431  
28480 0699-1318  
28480 0699-1391  
11502 W1206R03-2052B  
R418  
1
2
R420  
R421  
R422  
R423  
0699-2103  
0699-3431  
0699-1423  
0699-1431  
0699-1318  
0699-3763  
0699-2488  
0699-1433  
0699-1345  
0699-1826  
RESISTOR 49.9 +-1% .125 W TKF TC=0+-100  
RESISTOR 4.99K +-1% .125 W TKF TC=0+-100  
RESISTOR 215 +-1% .125 W TKF TC=0+-100  
RESISTOR 464 +-1% .125 W TKF TC=0+-100  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
RESISTOR 69 .1% 1206 .125 W 100V TC=25  
RESISTOR 100 +-0.1% .125 W TF TC=0+-25  
RESISTOR 562 +-1% .125 W TKF TC=0+-100  
RESISTOR 11 +-1% .125 W TKF TC=0+-100  
RESISTOR 45.3 +-1% .125 W TKF TC=0+-100  
28480 0699-2103  
28480 0699-3431  
28480 0699-1423  
28480 0699-1431  
28480 0699-1318  
11502 W1206-R03-69R0-B  
11502 W1206R031000BT  
28480 0699-1433  
28480 0699-1345  
28480 0699-1826  
1
R424  
R501-R502  
R521-R522  
R601  
R602-R609  
R610  
10  
10  
3
15  
2
R611-R614  
R615  
R616-R622  
R623  
R624-R625  
R626  
R627-R628  
0699-1415  
0699-1433  
0699-1345  
0699-1826  
0699-1415  
0699-2712  
0699-1415  
RESISTOR 100 +-1% .125 W TKF TC=0+-100  
RESISTOR 562 +-1% .125 W TKF TC=0+-100  
RESISTOR 11 +-1% .125 W TKF TC=0+-100  
RESISTOR 45.3 +-1% .125 W TKF TC=0+-100  
RESISTOR 100 +-1% .125 W TKF TC=0+-100  
RESISTOR 255 +-1% .125 W TKF TC=0+-100  
RESISTOR 100 +-1% .125 W TKF TC=0+-100  
28480 0699-1415  
28480 0699-1433  
28480 0699-1345  
28480 0699-1826  
28480 0699-1415  
28480 0699-2712  
28480 0699-1415  
7
2
119  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
33120-66521 – Main PC Assembly  
Reference  
Designator  
Agilent Part  
Number  
Mfr.  
Code  
Mfr. Part  
Number  
Qty  
Part Description  
R629  
R630  
R631  
R632-R633  
R634  
0699-2712  
0699-1415  
0699-1432  
0699-1384  
0699-1415  
0699-1432  
0699-1384  
0699-1344  
0699-1432  
0699-2631  
RESISTOR 255 +-1% .125 W TKF TC=0+-100  
RESISTOR 100 +-1% .125 W TKF TC=0+-10  
RESISTOR 511 +-1% .125 W TKF TC=0+-100  
RESISTOR 4.64K +-1% .125 W TKF TC=0+-100  
RESISTOR 100 +-1% .125 W TKF TC=0+-10  
RESISTOR 511 +-1% .125 W TKF TC=0+-100  
RESISTOR 4.64K +-1% .125 W TKF TC=0+-100  
RESISTOR 10 +-1% .125 W TKF TC=0+-100  
RESISTOR 511 +-1% .125 W TKF TC=0+-100  
RESISTOR 2K +-1% .125 W TKF TC=0+-100  
28480 0699-2712  
28480 0699-1415  
28480 0699-1432  
28480 0699-1384  
28480 0699-1415  
28480 0699-1432  
28480 0699-1384  
28480 0699-1344  
28480 0699-1432  
28480 0699-2631  
R635  
R636-R637  
R638-R639  
R640  
R641  
4
1
R642  
R643  
R644  
R645  
R646  
R647-R648  
R649  
R650  
R651  
0699-1432  
0699-2196  
0699-1415  
0699-1432  
0699-1352  
0699-1332  
0699-1392  
0699-1330  
0699-1392  
0699-1318  
RESISTOR 511 +-1% .125 W TKF TC=0+-100  
RESISTOR 78.7 1% 1206 .125 W 200 V TC=100 28480 0699-2196  
28480 0699-1432  
RESISTOR 100 +-1% .125 W TKF TC=0+-100  
RESISTOR 511 +-1% .125 W TKF TC=0+-100  
RESISTOR 21.5 +-1% .125 W TKF TC=0+-100  
RESISTOR 196K +-1% .125 W TKF TC=0+-100 28480 0699-1332  
RESISTOR 11K +-1% .125W TKF TC=0+-100 28480 0699-1392  
RESISTOR 100K +-1% .125 W TKF TC=0+-100 28480 0699-1330  
28480 0699-1415  
28480 0699-1432  
28480 0699-1352  
1
2
2
3
RESISTOR 11K +-1% .125 W TKF TC=0+-100  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
28480 0699-1392  
28480 0699-1318  
R652  
R653  
R654  
R655  
R656  
R657  
R658  
R659  
R660  
R661  
R662  
0699-2103  
0699-1432  
0699-2631  
0699-1415  
0699-1366  
0699-1360  
0699-1432  
0699-1318  
0699-1398  
0699-1330  
RESISTOR 49.9 +-1% .125 W TKF TC=0+-100  
RESISTOR 511 +-1% .125 W TKF TC=0+-100  
RESISTOR 2K +-1% .125 W TKF TC=0+-100  
RESISTOR 100 +-1% .125 W TKF TC=0+-100  
RESISTOR 82.5 +-1% .125 W TKF TC=0+-100  
RESISTOR 46.4 +-1% .125 W TKF TC=0+-100  
RESISTOR 511 +-1% .125 W TKF TC=0+-100  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
28480 0699-2103  
28480 0699-1432  
28480 0699-2631  
28480 0699-1415  
28480 0699-1366  
28480 0699-1360  
28480 0699-1432  
28480 0699-1318  
3
1
2
RESISTOR 21.5K +-1% .125 W TKF TC=0+-100 28480 0699-1398  
RESISTOR 100K +-1% .125 W TKF TC=0+-100 28480 0699-1330  
R663  
R664  
0699-1421  
0699-2883  
0699-2843  
0699-2489  
0699-2631  
0699-1422  
2100-4199  
0699-2883  
0699-1394  
0699-1366  
2
3
2
RESISTOR 178 +-1% .125 W TKF TC=0+-100  
RESISTOR 1.4K 1% 1206 .125 W 200V TC=100 28480 0699-2883  
28480 0699-1421  
R701-R702  
R703-R704  
R705-R706  
R707-R708  
R710  
R711  
R712  
R713  
RESISTOR 100K +-0.1% .125 W TF TC=0+-25  
RESISTOR 10K +-0.1% .125 W TF TC=0+-25  
RESISTOR 2K +-1% .125 W TKF TC=0+-100  
RESISTOR 196 +-1% .125 W TKF TC=0+-100  
RESISTOR-TRMR 100 20% TKF TOP-ADJ 1-TRN 32997 3314G-1-101E  
RESISTOR 1.4K 1% 1206 .125 W 200V TC=100 28480 0699-2883  
RESISTOR 14.7K +-1% .125 W TKF TC=0+-100 28480 0699-1394  
11502 W1206R031003BT  
11502 W1206R031002BT  
28480 0699-2631  
2
1
28480 0699-1422  
1
RESISTOR 82.5 +-1% .125 W TKF TC=0+-100  
28480 0699-1366  
R715-R716  
R717  
R718  
R719-R720  
R722  
R723  
0699-1415  
0699-2883  
0699-1351  
0699-1415  
0699-1366  
0699-1351  
RESISTOR 100 +-1% .125 W TKF TC=0+-100  
RESISTOR 1.4K 1% 1206 .125 W 200V TC=100 28480 0699-2883  
RESISTOR 19.6 +-1% .125 W TKF TC=0+-100  
RESISTOR 100 +-1% .125 W TKF TC=0+-100  
RESISTOR 82.5 +-1% .125 W TKF TC=0+-100  
RESISTOR 19.6 +-1% .125 W TKF TC=0+-100  
28480 0699-1415  
2
28480 0699-1351  
28480 0699-1415  
28480 0699-1366  
28480 0699-1351  
120  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
33120-66521 – Main PC Assembly  
Reference  
Designator  
Agilent Part  
Number  
Mfr.  
Code  
Mfr. Part  
Number  
Qty  
4
Part Description  
R724  
R725  
0699-1437  
0699-1415  
0699-2064  
0699-1415  
0699-1827  
0699-2890  
0699-2890  
0699-1378  
0699-1387  
0699-1426  
RESISTOR 825 +-1% .125 W TKF TC=0+-100  
RESISTOR 100 +-1% .125 W TKF TC=0+-100  
RESISTOR 6.8 +-5% .125 W TKF TC=0+-500  
RESISTOR 100 +-1% .125 W TKF TC=0+-100  
28480 0699-1437  
28480 0699-1415  
2M627 MCR18J  
28480 0699-1415  
R726-R727  
R728  
R729-R730  
R731-R732  
R734-R735  
R738  
2
2
4
RESISTOR 130 1% 1206PKG TC=100 200V 1/8W 28480 0699-1827  
RESISTOR 3.3 5% 1206 .125 W 200V TC=500  
RESISTOR 3.3 5% 1206 .125 W 200V TC=500  
RESISTOR 2.61K +-1% .125 W TKF TC=0+-100  
RESISTOR 6.81K +-1% .125 W TKF TC=0+-100  
RESISTOR 287 +-1% .125 W TKF TC=0+-100  
2M627 MCR18EZHJ3R3E  
2M627 MCR18EZHJ3R3E  
28480 0699-1378  
28480 0699-1387  
28480 0699-1426  
1
3
1
R739  
R740  
R741  
0699-1437  
0699-2488  
0699-1437  
0699-1387  
0699-2488  
0699-3019  
0699-3019  
0699-3019  
0699-3019  
0699-3022  
RESISTOR 825 +-1% .125 W TKF TC=0+-100  
RESISTOR 100 +-0.1% .125 W TF TC=0+-25  
RESISTOR 825 +-1% .125 W TKF TC=0+-100  
RESISTOR 6.81K +-1% .125 W TKF TC=0+-100  
RESISTOR 100 +-0.1% .125 W TF TC=0+-25  
RESISTOR 26.1 1% SM0805 .1 W TC=250 100V 28480 0699-3019  
RESISTOR 26.1 1% SM0805 .1 W TC=250 100V 28480 0699-3019  
RESISTOR 26.1 1% SM0805 .1 W TC=250 100V 28480 0699-3019  
RESISTOR 26.1 1% SM0805 .1 W TC=250 100V 28480 0699-3019  
28480 0699-1437  
11502 W1206R031000BT  
28480 0699-1437  
28480 0699-1387  
11502 W1206R031000BT  
R744-R747  
R748-R749  
R750-R751  
R752-R755  
R756  
14  
R758  
R760-R764  
R766-R772  
R774-R775  
2
4
RESISTOR 46.4 +-1% .1 W TKF TC=0+-200  
28480 0699-3022  
R801  
0699-3769  
0699-1423  
0699-3770  
0699-3765  
0699-3762  
0699-3766  
0699-3763  
0699-3767  
0699-3763  
0699-1423  
RESISTOR 11.8 +-1% .125 W TKF TC=0+-100  
RESISTOR 215 +-1% .125 W TKF TC=0+-100  
RESISTOR 23.7 +-0.1% .125 W TF TC=0+-25  
RESISTOR 115 .1% 1206 .125 W 100V TC=25  
RESISTOR 53 .1% 1206 .125 W 100V TC=25  
RESISTOR 232 .1% 1206 .125 W 100V TC=25  
RESISTOR 69 .1% 1206 .125 W 100V TC=25  
RESISTOR 309 .1% 1206 .125 W 100V TC=25  
RESISTOR 69 .1% 1206 .125 W 100V TC=25  
RESISTOR 215 +-1% .125 W TKF TC=0+-100  
28480 0699-3769  
28480 0699-1423  
R802-R803  
R804-R806  
R807-R808  
R809  
R810-R811  
R812-R813  
R814  
4
4
4
2
11502 W1206-R03-23R7-B  
11502 W1206-R03-1150-B  
11502 W1206-R03-53R0-B  
11502 W1206-R03-2320-B  
11502 W1206-R03-69R0-B  
11502 W1206-R03-3090-B  
11502 W1206-R03-69R0-B  
28480 0699-1423  
2
2
2
R815  
R816-R817  
R818  
R819  
R820  
R821  
R822-R823  
R824-R825  
R826-R827  
R828-R829  
R830-R832  
R833  
0699-2852  
0699-3770  
0699-2852  
0699-3769  
0699-3764  
0699-3765  
0699-3763  
0699-3769  
0699-3762  
0699-3767  
RESISTOR 442 +-1% .125 W TKF TC=0+-100  
RESISTOR 23.7 +-0.1% .125 W TF TC=0+-25  
RESISTOR 442 +-1% .125 W TKF TC=0+-100  
RESISTOR 11.8 +-1% .125 W TKF TC=0+-100  
28480 0699-2852  
11502 W1206-R03-23R7-B  
28480 0699-2852  
28480 0699-3769  
RESISTOR 442 0.1% 1206 .125W 100V TC=25 11502 W1206-R03-4420-B  
RESISTOR 115 .1% 1206 .125W 100V TC=25  
RESISTOR 69 .1% 1206 .125 W 100V TC=25  
RESISTOR 11.8 +-1% .125 W TKF TC=0+-100  
RESISTOR 53 .1% 1206 .125 W 100V TC=25  
RESISTOR 309 .1% 1206 .125 W 100V TC=25  
11502 W1206-R03-1150-B  
11502 W1206-R03-69R0-B  
28480 0699-3769  
11502 W1206-R03-53R0-B  
11502 W1206-R03-3090-B  
R834-R836  
R837  
R901  
R902  
R904-R905  
R906  
0699-3763  
0699-1383  
0699-1318  
0699-1330  
0699-1318  
0699-1374  
0699-1398  
RESISTOR 69 .1% 1206 .125 W 100V TC=25  
RESISTOR 4.22K +-1% .125 W TKF TC=0+-100  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
11502 W1206-R03-69R0-B  
28480 0699-1383  
28480 0699-1318  
7
1
RESISTOR 100K +-1% .125 W TKF TC=0+-100 28480 0699-1330  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
RESISTOR 1.78K +-1% .125 W TKF TC=0+-100  
RESISTOR 21.5K +-1% .125 W TKF TC=0+-100  
28480 0699-1318  
28480 0699-1374  
28480 0699-1398  
R907  
121  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
33120-66521 – Main PC Assembly  
Reference  
Designator  
Agilent Part  
Number  
Mfr.  
Code  
Mfr. Part  
Number  
Qty  
Part Description  
R908  
R909  
R910  
R1001  
R1002  
R1003  
R1004  
R1008-R1009  
R1010  
R1011  
0699-1318  
0699-3408  
0699-1391  
0699-1380  
0699-1424  
0699-1380  
0699-1424  
0699-1391  
0699-1421  
0699-1433  
RESISTOR 1K +-1% .125 W TKF TC=0+-100  
RESISTOR 1K +-5% 1 W TKF TC=0+-200  
RESISTOR 10K +-1% .125 W TKF TC=0+-100  
RESISTOR 3.16K +-1% .125 W TKF TC=0+-100 28480 0699-1380  
RESISTOR 237 +-1% .125 W TKF TC=0+-100 28480 0699-1424  
RESISTOR 3.16K +-1% .125 W TKF TC=0+-100 28480 0699-1380  
RESISTOR 237 +-1% .125 W TKF TC=0+-100  
RESISTOR 10K +-1% .125 W TKF TC=0+-100  
RESISTOR 178 +-1% .125 W TKF TC=0+-100  
RESISTOR 562 +-1% .125W TKF TC=0+-100  
28480 0699-1318  
2M627 MCR1001KJ  
28480 0699-1391  
1
3
2
28480 0699-1424  
28480 0699-1391  
28480 0699-1421  
28480 0699-1433  
R1012  
R1013  
R1014  
R1052  
0699-1391  
0699-1380  
0699-1384  
0699-1327  
RESISTOR 10K +-1% .125 W TKF TC=0+-100  
RESISTOR 3.16K +-1% .125 W TKF TC=0+-100 28480 0699-1380  
RESISTOR 4.64K +-1% .125 W TKF TC=0+-100 28480 0699-1384  
28480 0699-1391  
1
RESISTOR 1M +-1% .125 W TKF TC=0+-100  
28480 0699-1327  
RP301-RP302  
RP401  
RP403-RP404  
RP601  
1810-1360  
1810-1065  
1810-1331  
1810-1159  
2
1
2
1
RESISTOR-NETWORK 16PINS,THICK FILM,SMD 28480 1810-1360  
NET-RES 15 1.0K OHM 16-PIN  
NET-RES 8 220.0 OHM 16-PIN  
RESISTOR-NETWORK  
11236 767161102G  
11236 767163221G  
28480 1810-1159  
SHD902  
SPR1-SPR2  
SPR10-SPR20  
33120-00614  
0380-0643  
2190-0577  
1
2
2
RFI SHIELD  
STANDOFF-HEX .255-IN-LG 6-32-THD  
WASHER- NO. 10 .194-IN-ID .294-IN-OD  
28480 33120-00614  
28480 0380-0643  
20859 03118  
U101  
U102  
U103  
U104  
U105  
U106  
U107  
U108  
U109  
U201  
U202  
1990-1552  
1821-1479  
1820-8907  
33120-88861  
1818-4777  
1818-5699  
1820-5808  
1820-5941  
1820-5944  
1820-6306  
1820-5808  
3
1
1
1
1
1
2
1
2
3
OPTO-ISOLATOR LED-IC GATE IF=10 MA-MAX  
IC-BIT SLICE MPU/MCU  
IC GATE-ARY CMOS  
PROG PAL  
IC 256K-BIT SRAM 70-NS CMOS  
28480 HCPL-2211-300  
34649 N80C196KB  
27014 SCX6206AK0  
28480 33120-88861  
28480 1818-4777  
IC, CMOS, AMB, 32 PIN, 120 NANOSECONDS 34335 AM29F010-120JC  
IC FF CMOS/74HC D-TYPE POS-EDGE  
IC DCDR CMOS/74ACT BIN 8-TO-1-LINE  
74ACT32-GATE, QUAD 2-INPUT OR  
IC TRANSCEIVER CMOS/HCT BUS OCTL  
IC FF CMOS/74HC D-TYPE POS-EDGE  
04713 MC74HC273DW  
28480 1820-5941  
28480 1820-5944  
27014 MM74HCT245WM  
04713 MC74HC273DW  
U203-U204  
U205  
U206  
U210  
U211  
U212  
U213  
U214  
U215  
U216  
U217  
U218  
1820-6306  
1818-5093  
1821-0976  
1820-7244  
1820-5940  
1820-5944  
1990-1552  
1820-8830  
1820-5937  
1820-4998  
1820-5943  
1821-0559  
3
1
1
2
1
IC TRANSCEIVER CMOS/HCT BUS OCTL  
IC 256K-BIT SRAM 25-NS CMOS  
IC MODULATOR ANLG 68 PIN PLCC  
IC SCHMITT-TRIG CMOS/74AC INV HEX  
IC GATE CMOS/74ACT AND QUAD 2-INP  
74ACT32-GATE, QUAD 2-INPUT OR  
OPTO-ISOLATOR LED-IC GATE IF=10MA-MAX  
IC MUXR/DATA-SEL CMOS/ACT 8-TO-1-LINE  
IC FF CMOS/74AC D-TYPE POS-EDGE-TRIG  
IC GATE TTL/F OR QUAD 2-INP  
27014 MM74HCT245WM  
04713 MCM6206DJ25  
01295 F107563FN  
27014 74AC14SC  
28480 1820-5940  
07263 02237  
28480 HCPL-2211-300  
04713 MC74ACT251D  
07263 02237  
2
1
1
2
1
27014 74F32SC  
IC MUXR/DATA-SEL CMOS/74ACT 2-TO-1-LINE 07263 02237  
IC FF BICMOS/ABT D-TYPE POS-EDGE-TRIG 01295 SN74ABT16374ADL  
122  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
33120-66521 – Main PC Assembly  
Reference  
Designator  
Agilent Part  
Number  
Mfr.  
Code  
Mfr. Part  
Number  
Qty  
Part Description  
U301-U302  
U303  
U304  
U305  
U306-U307  
U314  
U315  
U401  
U402-U403  
U404-U405  
1821-0964  
1826-2793  
1821-0434  
1820-5790  
1826-1622  
1826-1622  
1820-5943  
1820-6524  
1820-8461  
1818-4983  
2
1
3
1
4
IC DRVR 8X S 20SOL 45V 250MA  
IC DA VOUT SER 16SOL 16BIT AD1851R  
IC ANLG-MUXR/DEMUXR CMOS/HC 8-CHAN  
01295 TPIC6595DW  
24355 AD1851R  
18324 74HC4051D  
IC SHF-RGTR CMOS/74HC SYNCHRO SERIAL-IN 18324 74HC4094D  
IC OP AMP LOW-BIAS-H-IMPD QUAD 14 PIN  
IC OP AMP LOW-BIAS-H-IMPD QUAD 14 PIN  
IC MUXR/DATA-SEL CMOS/74ACT 2-TO-1-LINE  
IC XLTR ECL/10KH TTL-TO-ECL QUAD  
IC FF BICMOS/ABT D-TYPE POS-EDGE-TRIG 01295 SN74ABT574ADW  
IC 256K-BIT SRAM 15-NS CMOS  
04713 TL074CD  
04713 TL074CD  
07263 02237  
1
2
2
04713 MC10H124FN  
28480 1818-4983  
U407  
U409  
U410  
U411  
U412  
U413  
U601-U602  
U603  
U604  
1826-3517  
1826-1619  
1826-1622  
1820-4377  
1820-5732  
1813-0861  
1821-0434  
1826-2797  
1820-8830  
1826-1950  
1
1
D/A 12-BIT 28-PLCC MISC  
D/A 8-BIT 20-PLCC CMOS  
IC OP AMP LOW-BIAS-H-IMPD QUAD 14 PIN  
IC GATE TTL/F NAND QUAD 2-INP  
IC LCH CMOS/74HC D-TYPE OCTL  
CLOCK-OSCILLATOR-XTAL 40.000-MHZ 0.01% 28480 1813-0861  
IC ANLG-MUXR/DEMUXR CMOS/HC 8-CHAN  
IC MULTIPLIER HS 14 PIN DIP-P  
IC MUXR/DATA-SEL CMOS/ACT 8-TO-1-LINE  
IC COMPARATOR HS SGL 8 PIN PLSTC-SOIC 28480 1826-1950  
28480 1826-3517  
24355 AD7528JP  
04713 TL074CD  
27014 74F00SC  
04713 MC74HC573ADW  
1
1
1
1
18324 74HC4051D  
24355 AD734AN  
04713 MC74ACT251D  
U620  
U621  
U702  
U901  
U902  
U903  
U904  
U906  
U907  
U908  
U1001  
1821-0622  
1826-1991  
1990-1552  
1820-7244  
34401-88842  
1821-1721  
1820-7662  
1820-6176  
1820-6175  
1826-0527  
1
1
IC DRVR BICMOS/ABT LINE OCTL  
18324 74ABT540D  
24355 AD711JR  
IC OP AMP HS SINGLE 8 PIN PLSTC-SOIC  
OPTO-ISOLATOR LED-IC GATE IF=10MA-MAX  
IC SCHMITT-TRIG CMOS/74AC INV HEX  
IC, ROM PROGRAMMED  
28480 HCPL-2211-300  
27014 74AC14SC  
28480 34401-88842  
01295 MP9914FNL  
28480 1820-7662  
01295 SN75ALS160DW  
01295 SN75ALS162DW  
27014 LM337T  
1
1
1
1
1
2
IC GPIB CONTROLLER  
IC-INTERFACE DRVR/RCVR BIPOLAR DUAL  
IC-INTERFACE XCVR BIPOLAR BUS OCTL  
IC-INTERFACE XCVR BIPOLAR BUS OCTL  
IC V RGLTR-ADJ-NEG 1.2/37V 3-TO-220 PKG  
U1002  
U1003  
U1004  
U1006  
U1007  
U1051  
U1052  
1826-0393  
1826-0527  
1826-1597  
1826-2264  
1826-2801  
1826-2794  
1826-2264  
1
IC V RGLTR-ADJ-POS 1.2/37V 3-TO-220 PKG  
IC V RGLTR-ADJ-NEG 1.2/37V 3-TO-220 PKG  
IC V RGLTR-FXD-POS 4.85/5.15V TO-220 PKG  
IC PWR MGT-UND-V-SEN 8 PINS P-SOIC PKG  
IC V RGLTR-OV-V-SEN 8-P-SOIC PKG  
27014 LM317  
27014 LM337T  
27014 LM2940CT  
04713 MC34064D-5  
04713 MC3423D  
27014 LM2490CT-5.0-LB01  
04713 MC34064D-5  
1
2
1
1
IC V RGLTR-FXD-POS 4.85/5.15V 3-TO-220  
IC PWR MGT-UND-V-SEN 8 PINS P-SOIC PKG  
XJ601  
XJ801  
3050-0447  
3050-0447  
2
2
2
WASHER-FL NM NO. 8 .192-IN-ID .37-IN-OD  
WASHER-FL NM NO. 8 .192-IN-ID .37-IN-OD  
28480 3050-0447  
28480 3050-0447  
XQ713  
XQ718  
1200-0181  
1200-0181  
INSULATOR-XSTR NYLON  
INSULATOR-XSTR NYLON  
13103 7717-5-N  
13103 7717-5-N  
7
Y101  
Y901  
0410-4009  
0410-4009  
CERO-RES 12 MHZ +1-0.8%  
CERO-RES 12 MHZ +1-0.8%  
28480 0410-4009  
28480 0410-4009  
123  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
33120-66502 – Display and Keyboard PC Assembly  
33120-66502 – Display and Keyboard PC Assembly  
Reference  
Designator  
Agilent Part  
Number  
Mfr.  
Code  
Mfr. Part  
Number  
Qty  
Part Description  
C1101  
C1102-1103  
C1104  
0160-5945  
0180-3751  
0160-6497  
0160-5947  
0160-6497  
0160-5947  
0160-5945  
0160-6497  
3
10  
3
CAP-FXD 0.01 uF 50 V  
CAP-FXD 1 uF 35 V TA  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 1000 pF 50 V  
CAP-FXD 0.1 uF 25 V  
CAP-FXD 1000 pF 50 V  
CAP-FXD 0.01 uF 50 V  
CAP-FXD 0.1 uF 25 V A  
04222 08055C103KAT A  
S0545 NRS105M35R8  
04222 12065C104KAT A  
04222 08055C102KAT A  
04222 12065C104KAT A  
04222 08055C102KAT A  
04222 08055C103KAT A  
04222 12065C104KAT A  
C1105  
C1201-1203  
C1301-1302  
C1303-1304  
C1305-1310  
CR1101  
CR1102  
CR1105  
J1301  
L1301  
M1301  
1906-0291  
1906-0395  
1902-1542  
34401-61602  
9170-1431  
33120-00611  
1
1
1
1
1
1
DIODE- 70 V 100 MA  
04713 MBAV99  
25403 BAS28  
04713 BZX84C6V2  
28480 34401-61602  
28480 9170-1431  
28480 33120-00611  
DIO SIG 2X SOT143L 75V 215MA 2PF  
DIODE-ZNR 6.2V 5% TO-236 (SOT-23)  
DISPLAY CABLE ASSY, 8.8L  
SHIELDING CORE, BEADS  
SHIELD-ESD  
R1101  
R1102  
R1103-1106  
R1107  
R1108  
R1109  
R1301-1304  
R1305-1306  
R1307-1308  
R1309  
0699-1399  
0699-1391  
0699-1344  
0699-1391  
0699-1378  
0699-1435  
0699-1423  
0699-1391  
0699-1330  
0699-1399  
2
4
4
RESISTOR 23.7K +-1% .125W TKF TC=0+-100  
RESISTOR 10K +-1% .125W TKF TC=0+-100  
RESISTOR 10 +-1% .125 W TKF TC=0+-100  
RESISTOR 10K +-1% .125W TKF TC=0+-100  
RESISTOR 2.61K +-1% .125W TKF TC=0+-100  
RESISTOR 681 +-1% .125W TKF TC=0+-100  
RESISTOR 215 +-1% .125W TKF TC=0+-100  
RESISTOR 10K +-1% .125W TKF TC=0+-100  
RESISTOR 100K +-1% .125W TKF TC=0+-100  
2M627 MCR18FX  
2M627 MCR18FX  
28480 0699-1344  
2M627 MCR18FX  
2M627 MCR18FX  
2M627 MCR18FX  
2M627 MCR18FX  
2M627 MCR18FX  
2M627 MCR18FX  
1
1
4
2
1
RESISTOR 23.7K +-1% .125W TKF TC=0+-100 2M627 MCR18FX  
S1119  
U1101  
U1102  
U1103  
U1201  
U1202  
U1203  
U1204  
U1301  
U1302  
Y1101  
0960-0892  
33120-88813  
1826-1402  
1826-2264  
1820-5562  
1820-4966  
1820-5330  
33120-89301  
1826-1528  
1820-6756  
0410-4009  
ROTARY ENCODER  
87C51 PROG  
28480 0960-0892  
1
1
1
1
1
1
1
1
1
1
28480 33120-88813  
04713 MC78L05ACD  
04713 MC34064D-5  
01295 SN74HC02D  
01295 SN74HC74D  
01295 SN75518FN  
28480 33120-89301  
IC-V RGLTR-FXD-POS 4.8/5.2V 8-P-SOIC PKG  
IC-PWR MGT-UND-V-SEN 8 PINS P-SOIC PKG  
74HC02-GATE, QUAD 2-INPUT NOR INPUT  
IC-FF CMOS/74HC D-TYPE POS-EDGE-TRIG  
IC-INTERFACE DRIVER BIPOLAR DISPLAY  
VACUUM FLUORESCENT DISPLAY  
IC-COMPARATOR LP QUAD 14 PIN PLSTC-SOIC 27014 LM339M  
IC-SHIFT-REGISTER CMOS/74HC BIDIR  
CERO-RES 12 MHZ +1-0.8%  
04713 MC74HC299D  
28480 0410-4009  
124  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
33120A Mainframe  
33120A Mainframe  
Reference  
Designator  
Agilent Part  
Number  
Mfr.  
Code  
Mfr. Part  
Number  
Qty  
Part Description  
A1  
A2  
B1  
33120-66521  
33120-66502  
3160-0847  
33120-61601  
RS232-61601  
33250-13603  
33120-84131  
2110-0458  
1
1
1
2
1
1
1
1
1
MAIN PC ASSEMBLY  
DISPLAY AND KEYBOARD PC ASSEMBLY  
FAN-TBAX 12V 5.3-CFM NOM  
CABLE BNC  
RS-232 CABLE  
AGILENT INTUILINK ARB SOFTWARE CD  
COVER  
28480 33120-66521  
28480 33120-66502  
11855 DFB0412L-SG  
28480 33120-61601  
28480 RS232-61601  
28480 33250-13603  
28480 33120-84131  
75915 239.500  
CBL1-CBL2  
CBL4  
CD1  
CVR1  
F1  
FUSE-500 MAT  
CHASSIS  
FRM1  
33120-80111  
28480 33120-80111  
HDW1  
HDW2  
HDW3  
HDW4  
KIT1  
KYC1  
KYC2  
KYC3  
0380-1820  
0535-0154  
3050-1547  
34401-88304  
34401-86010  
33120-87411  
34401-43711  
34401-45011  
1
1
1
1
1
1
1
1
SPACER, .438 IN LG, .280 IN OD  
NUT-HEX SGL-CHAM M9.0 X 0.75 2MM-THK  
WASHER-FL MTLC 9.0 9.2-MM-ID 14-MM-OD AL 11239 4-9-01  
REAR PANEL  
BUMPERS (FRNT/REAR) & POWER MOD CVR  
FRONT-PANEL KNOB  
PUSH-ROD, POWER-ON  
28480 0380-1820  
11239 3-9-03  
28480 34401-88304  
28480 34401-86010  
28480 33120-87411  
28480 34401-43711  
28480 33120-45011  
CARRYING HANDLE  
MP2  
PNL1  
PS1  
SCR1-SCR7  
SCR8-SCR9  
T1  
33120-81911  
33120-40211  
33120-86201  
0515-0433  
0624-0862  
9100-5090  
1
1
1
7
2
1
1
FRONT-PANEL KEYPAD  
28480 33120-81911  
28480 33120-40211  
28480 33120-86201  
28480 0515-0433  
FRONT-PANEL/BEZEL ASSEMBLY  
POWER-MODULE/FUSE & FUSE DRWR  
SCREW-M4 X 8MM LG PAN-HD  
SCREW-TPG .750-IN-LG PAN-HD-PHL STL,1022 93907 225-44395-890  
TRANSFORMER-PWR 100/120/220/240V  
FRONT DISPLAY WINDOW  
08807 8470B45  
28480 33120-49321  
WD1  
33120-49321  
7
125  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 7 Replaceable Parts  
Manufacturer’s List  
Manufacturer’s List  
Mfr Code  
Manufacturer’s Name  
Manufacturer’s Address  
Zip Code  
00779  
01295  
02113  
04222  
04713  
05971  
07263  
07933  
08807  
09021  
11236  
11239  
11502  
11855  
13103  
18324  
20859  
24355  
24444  
24226  
25403  
27014  
27264  
28480  
32997  
34335  
34649  
51406  
71744  
75915  
76381  
93907  
2L446  
2M627  
S0545  
S4217  
Amp Inc  
Texas Instruments Inc  
Coilcraft Inc  
AVX Corp  
Motorola Inc  
Harrisburg, PA, U.S.A.  
Dallas, TX, U.S.A.  
Cary, IL, U.S.A.  
Great Neck, NY, U.S.A.  
Roselle, IL, U.S.A.  
Burbank, CA, U.S.A.  
Cupertino, CA, U.S.A.  
Mountain View, CA, U.S.A.  
Grayslake, IL, U.S.A.  
Bradford, PA, U.S.A.  
Elkhart, IN, U.S.A.  
Alburquerque, NM, U.S.A.  
Corpus Christi, TX, U.S.A.  
Taipei, Taiwan  
17111  
75265  
60013  
11021  
60195  
91505  
95014  
94040  
60030  
16701  
46514  
87109  
78411  
LYN-TRON Inc  
Fairchild Semiconductor Corp  
Raytheon Co Semiconductor Div Hq  
MidwestCo Enterprises Inc  
KOA Speer Electronics Inc  
CTS Corp  
Nobel Mercantile Co  
IRC Inc  
Delta Electronic Industries Co  
Thermalloy Inc  
Dallas, TX, U.S.A.  
75234  
94086  
53212  
02062  
85281  
14070  
02876  
95052  
60532  
94303  
92507  
94086  
95054  
30067  
07012  
60016  
55144  
61101  
94063  
Signetics Corp  
Mellowes Co  
Sunnyvale, CA, U.S.A.  
Milwaukee, WI, U.S.A.  
Norwood, MA, U.S.A.  
Tempe, AZ, U.S.A.  
Gowanda, NY, U.S.A.  
Eindhoven, Netherlands  
Santa Clara, CA, U.S.A.  
Lisle, IL, U.S.A.  
Palo Alto, CA, U.S.A.  
Riverside, CA, U.S.A.  
Sunnyvale, CA, U.S.A.  
Santa Clara, CA, U.S.A.  
Marietta, GA, U.S.A.  
Clifton, NJ, U.S.A.  
Des Plaines, IL, U.S.A.  
St. Paul, MN, U.S.A  
Rockford, IL, U.S.A.  
Redwood City, CA, U.S.A.  
Kyoto 615, JAPAN  
Analog Devices Inc  
General Semiconductor Ind Inc  
Gowanda Electronics Corp  
NV Philips Elcoma  
National Semiconductor Corp  
Molex Inc  
Agilent Technologies, Inc.  
Bourns Networks Inc  
Advanced Micro Devices Inc  
Intel Corp  
Murata Corporation Of America  
General Instrument Corp  
Littelfuse Inc  
3M Corp  
Camcar Screw and Mfg Co  
ADAC Screw Machine Products  
Rohm Corp  
NEC Electronics Inc  
Nippon Chemi-Con Corp  
Mountain View, CA, U.S.A.  
Ohme-shi Tokyo, JAPAN  
94043  
126  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8
Schematics  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Schematics  
œ 33120A Block Diagram . . . . . . . . . . . . . . . . . 129  
œ Mechanical Disasembly . . . . . . . . . . . . . . . . . 130  
œ Floating Logic Schematic . . . . . . . . . . . . . . . . 131  
œ Digital Waveform Data Synthesis Schematic . . . . . 132  
œ System DAC Schematic . . . . . . . . . . . . . . . . . 133  
œ Waveform DAC Schematic . . . . . . . . . . . . . . . 134  
œ Filters Schematic . . . . . . . . . . . . . . . . . . . . 135  
œ Sync, Square Wave, and Attenuator Schematic . . . . 136  
œ Output Amplifier Schematic . . . . . . . . . . . . . . 137  
œ Output Attenuator Schematic . . . . . . . . . . . . . 138  
œ Earth Reference Logic Schematic . . . . . . . . . . . . 139  
œ Power Supplies Schematic . . . . . . . . . . . . . . . 140  
œ Display and Keyboard Schematic . . . . . . . . . . . . 141  
œ 33120-66521 Component Locator Diagram . . . . . . 142  
œ 33120-66502 Component Locator Diagram . . . . . . 143  
You may notice parts labeled as “No Load” on several schematics.  
These are parts that were included in the original design but were  
removed later to enhance performance or reduce cost.  
128  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Block Diagram  
129  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Loosen  
Captive  
Screws  
Pull Off Handle  
Slide Off Outer Case  
Pull Off Bumpers  
Remove Bottom Screw  
Pull To  
Remove  
Pry Outward  
From This Side  
Press Down To  
Unlatch Cable  
Pull Up To  
Disconnect  
Remove Screw  
Front Panel  
Will Pull Off  
Mechanical Disassembly  
130  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33120-66521 (sheet 1 of 10)  
Floating Logic Schematic  
131  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33120-66521 (sheet 2 of 10)  
Digital Waveform Data Synthesis Schematic  
132  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33120-66521 (sheet 3 of 10)  
System DAC Schematic  
133  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33120-66521 (sheet 4 of 10)  
Waveform DAC Schematic  
134  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33120-66521 (sheet 5 of 10)  
Filters Schematic  
135  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33120-66521 (sheet 6 of 10)  
Sync, Square Wave, and Attenuator Schematic  
136  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33120-66521 (sheet 7 of 10)  
Output Amplifier Schematic  
137  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33120-66521 (sheet 8 of 10)  
Output Attenuator Schematic  
138  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33120-66521 (sheet 9 of 10)  
Earth Reference Logic Schematic  
139  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33120-66521 (sheet 10 of 10)  
Power Supplies Schematic  
140  
Download from Www.Somanuals.com. All Manuals Search And Download.  
33120-66502 (sheet 1 of 1)  
Display and Keyboard Schematic  
141  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TOP SIDE 33120-66521 REV A  
33120-66521 Component Locator Diagram  
142  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TOP SIDE 33120-66502 REV C  
33120-66502 Component Locator Diagram  
143  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Copyright Agilent Technologies, Inc.  
Warranty  
Safety Notices  
1994-2002  
The material contained in this  
document is provided “as is,” and  
is subject to being changed, with-  
out notice, in future editions.  
Further, to the maximum extent  
permitted by applicable law,  
Agilent disclaims all warranties,  
either express or implied with  
regard to this manual and any  
information contained herein,  
including but not limited to the  
implied warranties of merchant-  
ability and fitness for a particular  
purpose. Agilent shall not be  
liable for errors or for incidental  
or consequential damages in  
connection with the furnishing,  
use, or performance of this docu-  
ment or any information contain-  
ed herein. Should Agilent and the  
user have a separate written  
agreement with warranty terms  
covering the material in this docu-  
ment that conflict with these  
terms, the warranty terms in the  
separate agreement will control.  
Do not install substitute parts or  
perform any unauthorized modifica-  
tion to the product. Return the  
product to an Agilent Technologies  
Sales and Service Office for service  
and repair to ensure that safety  
features are maintained.  
No part of this manual may be repro-  
duced in any form or by any means  
(including electronic storage and  
retrieval or translation into a foreign  
language) without prior agreement  
and written consent from Agilent  
Technologies as governed by the  
United States and international  
copyright laws.  
W A R N I N G  
A WARNING notice denotes a hazard.  
It calls attention to an operating proce-  
dure, practice, or the like that, if not  
correctly performed or adhered to,  
could result in personal injury or death.  
Do not proceed beyond a WARNING  
notice until the indicated conditions  
are fully understood and met.  
Manual Part Number  
33120-90017, March 2002  
(order as 33120-90104 manual set)  
Edition  
Edition 6, March 2002  
Edition 5, August 1997  
Edition 4, February 1996  
Edition 3, May 1994  
Edition 2, March 1994  
Edition 1, January 1994  
C A U T I O N  
A CAUTION notice denotes a hazard.  
It calls attention to an operating proce-  
dure, practice, or the like that, if not  
correctly performed or adhered to,  
could result in damage to the product  
or loss of important data. Do not  
proceed beyond a CAUTION notice  
until the indicated conditions are  
fully understood and met.  
Printed in Malaysia  
Agilent Technologies, Inc.  
815 14th Street S.W.  
Loveland, Colorado 80537 U.S.A.  
Technologies Licenses  
The hardware and/or software  
described in this document are  
furnished under a license and may  
be used or copied only in accordance  
with the terms of such license.  
Assistance  
Product maintenance agreements and  
other customer assistance agreements  
are available for Agilent Technologies  
products. For assistance, contact your  
nearest Agilent Technologies Sales  
and Service Office. Further information  
is available on the Agilent web site at  
www.agilent.com/find/assist.  
Restricted Rights Legend  
Earth ground symbol.  
If software is for use in the perform-  
ance of a U.S. Government prime  
contract or subcontract, Software is  
delivered and licensed as “Commercial  
computer software” as defined in  
DFAR 252.227-7014 (June 1995),  
or as a “commercial item” as defined  
in FAR 2.101(a) or as “Restricted  
computer software” as defined in  
FAR 52.227-19 (June 1987) or any  
equivalent agency regulation or con-  
tract clause. Use, duplication or  
disclosure of Software is subject to  
Agilent Technologies’ standard  
Chassis ground symbol.  
Trademark Information  
Microsoft® and Windows® are U.S.  
registered trademarks of Microsoft  
Corporation. All other brand and  
product names are trademarks or  
registered trademarks of their  
respective companies.  
W A R N I N G  
Only qualified, service-trained person-  
nel who are aware of the hazards  
involved should remove the cover from  
the instrument.  
commercial license terms, and non-  
DOD Departments and Agencies of  
the U.S. Government will receive no  
greater than Restricted Rights as  
defined in FAR 52.227-19(c)(1-2)  
(June 1987). U.S. Government users  
will receive no greater than Limited  
Rights as defined in FAR 52.227-14  
(June 1987) or DFAR 252.227-7015  
(b)(2) (November 1995), as applicable  
in any technical data.  
Certification  
W A R N I N G  
Agilent Technologies certifies that  
this product met its published specifi-  
cations at the time of shipment from  
the factory. Agilent Technologies  
further certifies that its calibration  
measurements are traceable to the  
United States National Institute of  
Standards and Technology, to the  
extent allowed by that organization’s  
calibration facility, and to the calibra-  
tion facilities of other International  
Standards Organization members.  
For continued protection against fire,  
replace the line fuse only with a fuse  
of the specified type and rating.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

3Com Digital Camera 10242063 00 User Manual
3D Connexionr S201 User Manual
Adaptec Network Card Storage Manager User Manual
Agri Fab Chipper 2468 User Manual
Alliance Laundry Systems Treadmill 815 User Manual
Alto Shaam Food Warmer PD2SYS 48 User Manual
AOSafety Respiratory Product Half Mask Respirator User Manual
Atlas Indoor Furnishings MLG180B User Manual
Bakers Pride Oven Fryer FKM User Manual
Eureka Tents Tent Nergy Portable Canopy User Manual