Agilent Technologies Power Supply 6625A User Manual

OPERATING MANUAL  
MULTIPLE OUTPUT LINEAR SYSTEM  
DC POWER SUPPLY  
AGILENT MODELS 6625A, 6626A,  
6628A, and 6629A  
Agilent Part No 06626-90001  
Microfiche Part No. 06626-90002  
Printed in Malaysia  
September, 2004  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SAFETY SUMMARY  
The following general safety precautions must be observed during all phases of operation, service, and repair of this  
instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety  
standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the  
customer’s failure to comply with these requirements.  
BEFORE APPLYING POWER.  
Verify that the product is set to match the available line voltage and the correct fuse is installed.  
GROUND THE INSTRUMENT.  
This product is a Safety Class 1 instrument (provided with a protective earth terminal). To minimize shock hazard, the instrument chassis  
and cabinet must be connected to an electrical ground. The instrument must be connected to the ac power supply mains through a three-  
conductor power cable, with the third wire firmly connected to an electrical ground (safety ground) at the power outlet. For instruments  
designed to be hard-wired to the ac power lines (supply mains), connect the protective earth terminal to a protective conductor before any  
other connection is made. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will  
cause a potential shock hazard that could result in personal injury. If the instrument is to be energized via an external autotransformer for  
voltage reduction, be certain that the autotransformer common terminal is connected to the neutral (earthed pole) of the ac power lines  
(supply mains).  
FUSES.  
Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired  
fuses or short circuited fuseholders. To do so could cause a shock or fire hazard.  
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE.  
Do not operate the instrument in the presence of flammable gases or fumes.  
KEEP AWAY FROM LIVE CIRCUITS.  
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified  
service personnel. Do not replace components with power cable connected. Under certain conditions, dangerous voltages may exist even  
with the power cable removed. To avoid injuries, always disconnect power, discharge circuits and remove external voltage sources before  
touching components.  
DO NOT SERVICE OR ADJUST ALONE.  
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.  
DO NOT EXCEED INPUT RATINGS.  
This instrument may be equipped with a line filter to reduce electromagnetic interference and must be connected to a properly grounded  
receptacle to minimize electric shock hazard. Operation at line voltages or frequencies in excess of those stated on the data plate may  
cause leakage currents in excess of 5.0 mA peak.  
SAFETY SYMBOLS.  
Instruction manual symbol: the product will be marked with this symbol when it is necessary for the user to refer to the  
instruction manual (refer to Table of Contents) .  
Indicates hazardous voltages.  
Indicate earth (ground) terminal.  
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly  
performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the  
indicated conditions are fully understood and met.  
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like, which, if not correctly  
performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed  
beyond a CAUTION sign until the indicated conditions are fully understood and met.  
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT.  
Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the  
instrument. Return the instrument to an Agilent Technologies Sales and Service Office for service and repair to ensure that safety features  
are maintained.  
Instruments which appear damaged or defective should be made inoperative and secured against unintended operation until they can be  
repaired by qualified service personnel  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
SAFETY SUMMARY (continued)  
GENERAL  
Any LEDs used in this product are Class 1 LEDs as per IEC 825-1.  
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 du Canada.  
ENVIRONMENTAL CONDITIONS  
This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to  
operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the specifications tables for the  
ac mains voltage requirements and ambient operating temperature range.  
SAFETY SYMBOL DEFINITIONS  
Symbol  
Description  
Direct current  
Symbol  
Description  
Terminal for Line conductor on permanently  
installed equipment  
Alternating current  
Caution, risk of electric shock  
Caution, hot surface  
Both direct and alternating current  
Three-phase alternating current  
Earth (ground) terminal  
Caution (refer to accompanying documents)  
In position of a bi-stable push control  
Protective earth (ground) terminal  
Frame or chassis terminal  
Out position of a bi-stable push control  
On (supply)  
Terminal for Neutral conductor on  
permanently installed equipment  
Off (supply)  
Terminal is at earth potential  
Standby (supply)  
(Used for measurement and control  
circuits designed to be operated with  
one terminal at earth potential.)  
Units with this symbol are not completely  
disconnected from ac mains when this switch is  
off. To completely disconnect the unit from ac  
mains, either disconnect the power cord or have  
a qualified electrician install an external switch.  
Herstellerbescheinigung  
Diese Information steht im Zusammenhang mit den Anforderungen der  
Maschinenläminformationsverordnung vom 18 Januar 1991.  
* Schalldruckpegel Lp <70 dB(A) * Am Arbeitsplatz * Normaler Betrieb  
* Nach EN 27779 (Typprüfung).  
Manufacturer’s Declaration  
This statement is provided to comply with the requirements of the German Sound Emission Directive,  
from 18 January 1991.  
* Sound Pressure Lp <70 dB(A) * At Operator Position * Normal Operation  
* According to EN 27779 (Type Test).  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
DECLARATION OF CONFORMITY  
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014  
Manufacturer’s Name and Address  
Responsible Party  
Alternate Manufacturing Site  
Agilent Technologies, Inc.  
550 Clark Drive, Suite 101  
Budd Lake, New Jersey 07828  
USA  
Agilent Technologies (Malaysia) Sdn. Bhd  
Malaysia Manufacturing  
Bayan Lepas Free Industrial Zone, PH III  
11900 Penang,  
Malaysia  
Declares under sole responsibility that the product as originally delivered  
Product Names  
a) Multiple Output 40 W and 80 W system dc Power Supplies  
b) Multiple Output 25 W and 50 W precision system dc Power Supplies  
Model Numbers  
a) 6621A, 6622A, 6623A; 6624A, 6627A  
b) 6625A; 6626A; 6628A; 6629A  
(and other customized products based upon the above)  
Product Options  
This declaration covers all options and customized products based on the above  
products.  
Complies with the essential requirements of the Low Voltage Directive 73/23/EEC and the EMC  
Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly.  
EMC Information  
ISM Group 1 Class A Emissions  
As detailed in  
Electromagnetic Compatibility (EMC), Certificate of Conformance Number  
CC/TCF/00/076 based on Technical Construction File (TCF) HPNJ3, dated  
Oct. 29, 1997  
Assessed by:  
Celestica Ltd, Appointed Competent Body  
Westfields House, West Avenue  
Kidsgrove, Stoke-on-Trent  
Straffordshire, ST7 1TL  
United Kingdom  
Safety Information  
and Conforms to the following safety standards.  
IEC 61010-1:2001 / EN 61010-1:2001  
UL 1244  
CSA C22.2 No. 1010.1:1992  
This DoC applies to the above-listed products placed on the EU market after:  
January 1, 2004  
Date  
Bill Darcy/ Regulations Manager  
For further information, please contact your local Agilent Technologies sales office, agent or distributor.  
Authorized EU-representative: Agilent Technologies Deutschland GmbH, Herrenberger Strabe 130, D  
71034 Böblingen, Germany  
Revision: B.00.00  
Issue Date: Created on 11/24/2003 3:33  
PM  
Document No. 662xA.11.24.doc  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
WHAT THIS MANUAL CONTAINS  
It contains information relating to the installation, operation, and programming of these supplies as outlined below.  
Maintenance and troubleshooting instructions are given in a separate Service Manual (Agilent Part No. 06626-90003).  
Chapter 1.--General Information  
Chapter 1 contains a general description of the power supplies as well as instrument specifications and information  
concerning options and accessories.  
Chapter 2.--Installation Procedures  
Chapter 2 contains information to prepare the supply for use. Included in this chapter are power requirements, line voltage  
conversion, and GP-IB interface connections.  
Chapter 3.--Getting Started  
Chapter 3 contains a brief description of the supplys front panel controls and indicators and describes how to turn on the  
supply and to check its operation. An introduction to remote operation over the GP-IB is also given to help a first time user  
get started quickly.  
Chapter 4.--Output Connections and Operating Information  
Chapter 4 contains information about making connections to the supplys output terminals. General operating information is  
also provided.  
Chapter 5.--Remote Operation  
Chapter 5 contains all of the information required to operate the supply remotely via a GP-IB computer. All of the  
commands that can be used to program the supplies are described.  
Chapter 6.--Local Operation  
Chapter 6 contains instructions on using all of the front panel controls and indicators.  
Appendix A--Calibration Procedure  
Appendix A contains programming steps and procedures that are required to calibrate your power supply. It is  
recommended that the power supply be calibrated yearly.  
Appendix B--Programming with Series 200/300 Computer  
Appendix B contains Series 200/300 Computer programming examples (in Agilent extended BASIC language) for your  
Power Supplys most frequently used functions.  
Appendix C--Command Summary  
Appendix C contains an alphabetical listing of all commands that can be sent to a supply.  
Appendix D--Error Messages  
Appendix D contains a listing and brief explanation of all error codes and messages for all programming and hardware  
errors.  
Appendix E - Manual Backdating  
Appendix E contains backdating information for units with Serial numbers lower than those listed on the title page.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table Of Contents  
1
General Information  
Introduction............................ .......................................................................................................................11  
Safety Considerations....................................................................................................................................11  
Instrument and Manual Identification.... .......................................................................................................11  
Options................................. .........................................................................................................................11  
Accessories....................................................................................................................................................12  
Description ............................ .......................................................................................................................12  
Basic Operation.............................................................................................................................................14  
GP-IB Board..............................................................................................................................................14  
Output Boards............................................................................................................................................15  
Specifications......................... .......................................................................................................................15  
Qualifying Conditions.................. .............................................................................................................15  
Definitions.................................................................................................................................................15  
2
Installation  
Introduction............................ .......................................................................................................................25  
Initial Inspection............................................................................................................................................25  
Location and Cooling.................... ................................................................................................................25  
Input Power Requirements................ ............................................................................................................26  
Line Fuse.......................................................................................................................................................27  
Power Cord....................................................................................................................................................28  
Line Voltage Conversion...............................................................................................................................29  
GP-IB Interface Connection..........................................................................................................................29  
3
Getting Started  
Introduction ........................... .......................................................................................................................31  
Front Panel Controls and Indicators..............................................................................................................31  
Turning On Your Supply...............................................................................................................................36  
Normal Self Test Indications.....................................................................................................................36  
Self Test Errors..........................................................................................................................................37  
Checking Out Your Supply Using Local Control..........................................................................................37  
Voltage Test............................ ..................................................................................................................38  
Overvoltage Test......................... ..............................................................................................................38  
Current Test...............................................................................................................................................38  
Changing Resolution Range......................................................................................................................39  
Introduction To Remote Operation....... ........................................................................................................39  
Enter/Output Statements............................................................................................................................40  
Reading the GP-IB Address.............. ........................................................................................................40  
Changing the GP-IB Address....................................................................................................................40  
Sending a Remote Command....................................................................................................................41  
Getting Data from the Supply....................................................................................................................41  
Often Used Commands..............................................................................................................................41  
Returning the Supply to Local Mode..... ...................................................................................................43  
4
Output Connections and Operating Information  
Introduction ............................... ...................................................................................................................45  
Output Ranges...............................................................................................................................................45  
Operating Quadrants..................................................................................................................................46  
Operating Range........................................................................................................................................47  
Protection Features........................................................................................................................................48  
Connecting The Load......................... ...........................................................................................................49  
Wire Size Selection......................... ..........................................................................................................49  
Multiple Loads.............................. ............................................................................................................50  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table Of Contents (continued)  
Positive and Negative Voltages.................................................................................................................51  
Remote Voltage Sensing..................... ..........................................................................................................51  
Remote Sense Connections................... ....................................................................................................52  
Output Noise Considerations.....................................................................................................................53  
Programming Response Time with an Output Capacitor............................. .............................................53  
Open Sense Leads......................................................................................................................................54  
Overvoltage Trigger Connections............ .....................................................................................................54  
External Trigger Circuit................... .........................................................................................................54  
Power Supply Protection Considerations .......................... ...........................................................................55  
Battery Charging........................... ............................................................................................................55  
Capacitive Load Limitation.......................................................................................................................56  
Parallel Operation..........................................................................................................................................56  
CV Operation............................... .............................................................................................................57  
CC Operation.............................................................................................................................................58  
Remote Sensing.........................................................................................................................................58  
Specifications for Parallel Operation.........................................................................................................58  
Series Operation....................... .....................................................................................................................59  
CV Operation............................... .............................................................................................................59  
CC Operation.............................................................................................................................................60  
Remote Sensing.........................................................................................................................................60  
Specifications for Series Operation...........................................................................................................60  
Bi-Polar Operation ....................................................................................................................................61  
5
Remote Operation  
Introduction ............................. .....................................................................................................................63  
GP-IB Operation............................................................................................................................................63  
Interface Functions....................................................................................................................................63  
GP-IB Address Selection...........................................................................................................................64  
Power-On Service Request (PON)............................................................................................................65  
Programming Syntax.....................................................................................................................................65  
Numeric Data............................... .............................................................................................................67  
Order of Execution......................... ...........................................................................................................70  
Terminators................................ ...............................................................................................................70  
Initial Conditions...........................................................................................................................................71  
Power Supply Commands .............................................................................................................................71  
Voltage Programming........................ .......................................................................................................71  
Current Programming................................................................................................................................73  
Range Programming..................................................................................................................................73  
Output On/Off.................................... .......................................................................................................74  
Overvoltage Protection..............................................................................................................................74  
Overcurrent Protection........................... ...................................................................................................75  
Multiple Output Storage and Recall..........................................................................................................75  
The Clear Command..................................................................................................................................76  
Status Reporting................................. .......................................................................................................76  
Service Request Generation....................... ...............................................................................................79  
Reprogramming Delay............................. .................................................................................................80  
Display On/Off..........................................................................................................................................81  
Other Queries.............................................................................................................................................82  
6
Local Operation  
Introduction ..................................... .............................................................................................................85  
Local Mode....................................................................................................................................................85  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table Of Contents (continued)  
Local Control Of Output Functions...............................................................................................................85  
General........................................... ...........................................................................................................85  
Setting Voltage and Voltage Range...........................................................................................................86  
Setting Current and Current Range............................................................................................................87  
Enabling/Disabling an Output...................... .............................................................................................87  
Setting Overvoltage Protection..................................................................................................................88  
Resetting Overvoltage Protection..............................................................................................................88  
Enabling/Disabling Overcurrent Protection........ ......................................................................................88  
Resetting Overcurrent Protection................. .............................................................................................88  
Local Control Of System Functions..............................................................................................................88  
Setting the Supplys GP-IB Address..........................................................................................................88  
Displaying Error Messages........................................................................................................................89  
Storing and Recalling Voltage and Current. Settings for all Outputs........................................................89  
A
B
Alignment Procedures  
Introduction...................................... .............................................................................................................91  
Test Equipment and Setup Required................. ............................................................................................91  
Calibration Program.............................. ........................................................................................................91  
Alignment Program.............................. .........................................................................................................95  
Programming With a Series 200/300 Computer  
Introduction ....................... .........................................................................................................................103  
I/O Path Names...................... .....................................................................................................................103  
Voltage and Current Programming..... ........................................................................................................103  
Voltage and Current Programming With Variables... .................................................................................104  
Voltage and Current Readback....................................................................................................................104  
Programming Power Supply Registers........................................................................................................105  
Present Status...................... ....................................................................................................................105  
Service Request and Serial Poll...............................................................................................................106  
Error Detection .......................................................................................................................................107  
Stored Operating States...........................................................................................................................108  
C
D
Command Summary  
Introduction.................................................................................................................................................109  
Error Messages  
Introduction....................... ..........................................................................................................................117  
Power-On Self Test Messages.....................................................................................................................117  
Error Responses...........................................................................................................................................117  
Test Responses...................... ......................................................................................................................117  
E
Manual Backdating  
Introduction ........................ ........................................................................................................................121  
Make Changes.................…………………………………………………………………………………...121  
Agilent Sales and Support Office  
Contacts ......................................................................................................................................................122  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1
General Information  
Introduction  
This chapter contains a general description of your power supply as well as its performance specifications. Information  
about options, accessories, and GP-IB cables is also provided. This manual describes the Agilent 6625A, 6626A, 6628A,  
and Agilent 6629A power supplies. Unless stated otherwise, the information in this manual applies to all of these models.  
Information that is specific to one model only is identified as such in this manual.  
Safety Considerations  
This product is a Safety Class 1 instrument, which means that it is provided with a protective earth terminal. This terminal  
must be connected to a power source that has a 3-wire ground receptacle. Review the instrument and this manual for safety  
markings and instructions before operation. Refer to the Safety Summary page at the beginning of this manual for a  
summary of general safety information. Safety information for specific procedures is located at appropriate places in this  
manual.  
Instrument and Manual Identification  
Agilent Technologies power supplies are identified by a two-part serial number, i.e. 2601A-00101. The first part of the  
serial number (the prefix) is a number/letter combination that denotes either the date of manufacture or the date of a  
significant design change. It also indicates the country of origin. (Starting at 1960, 26 = 1986; 01 = the first week of the  
year; A = U.S.A.) The second part of the serial number is a different sequential number assigned to each instrument starting  
with 00101.  
Options  
Options 100,120, 220, and 240 simply determine which line voltage is selected at the factory. For information about  
changing the line voltage setting, see line voltage conversion paragraph on page 29.  
Option 750 consists of a fault indicator (FLT) and remote inhibit (INH) circuit and relay control, which provide additional  
shutdown protection should either the GP-IB and/or controller fail. This Option is described in a separate document  
entitled, Appendix E Option 750 Operating Instructions for the Multiple Output Linear System DC Power Supply, Agilent  
Models 6621A, 6622A, 6623A, 6624A, and 6627A(Agilent P/N 5957-6372).  
#100 Input power, 100 Vac, 47-66 Hz  
#120 Input power, 120 Vac, 47-66 Hz  
#220 Input power, 220 Vac, 47-66 Hz  
#240 Input power, 240 Vac, 47-66 Hz  
#750 Fault (FLT) Remote Inhibit (INH) and Relay Control  
#908 One rack mount kit (5062-3977)  
#909 One rack mount kit with handles (5062-3983)  
#0L2 One extra operating manual  
#0B3 One service manual  
General Information 11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Accessories  
10833A GP-IB cable, 1 m (3.3 ft)  
10833B GP-IB cable, 2 m (6.6 ft)  
10833C GP-IB cable, 4 m (13.2 ft)  
10833D GP-IB cable, 0.5 m (1.6 ft)  
10834A GP-IB connector extender  
Slide mount kit (1494-0059)  
Description  
The Agilent power supply features a combination of programming capabilities and linear power supply performance that  
make them ideal for systems applications. The models in this family offer a total of up to 200 watts of output power, with  
voltages up to 50 volts and currents up to 2 amps. The output combinations that correspond to each model are shown in  
Table 1-1. Each isolated output has a high resolution and standard resolution range, as shown in Figure 1-1.  
Table 1-1 Output Combinations Available  
Model  
Output 1  
25 W  
25 W  
50 W  
50 W  
Output 2  
50 W  
25 W  
50 W  
50 W  
Output 3  
--  
50 W  
--  
Output 4  
Agilent 6625A  
Agilent 6626A  
Agilent 6628A  
Agilent 6629A  
--  
50 W  
--  
50 W  
50W  
Figure 1-1. Output Operating Range  
Output 2 of the Agilent 6625A and outputs 3 and 4 of the Agilent 6626A and all outputs of the Agilent 6628A and Agilent  
6629A have an "Lshaped characteristic. The L shaped characteristic allows you to use the same output to power loads  
with different voltage and current requirements. The power supply automatically selects one of the operating ranges based  
on the last parameter (voltage or current) that is set. If a high resolution or a standard resolution application is required for  
either voltage or current, VRSET and IRSET should be used (see Appendix C-Command Summary). Additionally, each  
output contains an active downprogrammer, which means that voltage downprogramming can be accomplished as quickly  
as upprogramming, even without a load.  
12 General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The power supplies allow up to a 10 volt sense lead drop. This feature makes them ideal for test system applications where  
remote sensing is used.  
The output voltage and current for any output can be monitored with the front panel display. Output specific error messages  
are also displayed. Front panel annunciators show the operating status of the instrument. The front panel keypad lets you set  
and readback the voltage limit, current limit, and overvoltage trip level of any output. With the keypad, you can also enable  
or disable outputs, enable overcurrent protection, reset overvoltage and overcurrent protection, and return to local operating  
mode.  
Your multiple output power supply can be both a listener and a talker on the GP-IB. (GP-IB is Agilent Technologies’  
implementation of IEEE-488). The built-in interface is tailored to the supply, resulting in simpler programming. Voltage  
and current settings can be sent directly to the specified output in volts and amps.  
Service can be requested from your power supply for up to 10 reasons. The supply responds to a serial poll by identifying  
the output on which the fault occurred. Self-contained measurement and readback capability eliminate the need for  
externally scanning the outputs using a separate DVM. Upon command, the supply will measure its output voltage or  
current and return the value on the GP-IB. The following functions are implemented via the GP-IB:  
Voltage and current programming.  
Programmable resolution range.  
Voltage and current measurement and readback.  
Present and accumulated status readback.  
Programmable service request mask.  
Programmable overvoltage and overcurrent protection.  
Voltage and current range programming.  
Storage and recall of programmed voltage and current values for all outputs.  
Queries of programmed functions or settings.  
Output enable or disable.  
Programming syntax error detection.  
Programmable delay time for service request and OCP mask.  
Voltage, current, and overvoltage calibration.  
GP-IB interface selftest.  
Message display capability on the front panel.  
Output connections are made to rear panel screw terminals. Either the positive or negative output terminal can be grounded,  
or the output can be floated up to ± 240 Vdc (including output voltage) from chassis ground. Output voltage can be locally  
or remotely sensed, and identical outputs can be operated in series or parallel combinations for increased output voltage or  
current capability. As shipped from the factory, the power supply is jumpered for local sensing.  
Your power supply can be calibrated without having to remove the cover or even having to remove it from your system  
cabinet. This feature allows you to calibrate the supply at its normal operating temperature. The recommended calibration  
interval is one year. Refer to Appendix A of this manual for complete calibration details. A calibration security jumper is  
available inside the unit. Access is described in the service manual.  
General Information 13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Basic Operation  
Figure 1-2 is a block diagram that illustrates the major assemblies contained within the power supply. As shown in the  
figure, each supply includes a power transformer, two or more output boards, a GP-IB board, and front panel (display and  
control keys).  
Figure 1-2. Block Diagram  
The appropriate ac input voltage is applied to each output board where it is converted to a raw dc voltage which is  
subsequently linearly regulated to become the dc output voltage. The magnitude of the output and the mode of operation are  
determined by the load and the data received from the GP-IB computer or from the front panel.  
Each power supply model contains one output board for each output that it provides.  
GP-IB Board  
The GP-IB board provides the interface between the user and the multiple outputs of the power supply. Each output board  
is actually an output channel that can be individually selected and controlled over the GP-IB or from the supplys front  
panel. Circuits on the GP-IB board interpret commands from the GP-IB or from the front panel to control the selected  
output.  
The GP-IB board also processes measurement and status data received from the output boards. This data may be read back  
over the GP-IB and/or displayed on the supplys front panel.  
The power supply has no potentiometers. Each output is individually calibrated over the GP-IB using calibration commands  
(see Appendix A). Correction factors are calculated by the power supply during calibration and are stored in a non-volatile  
memory which is located on the supplys GP-IB board. The supply contains no batteries.  
14 General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Output Boards  
The output boards are linear dc power supplies. Each isolated output operating boundary curve is shown in Figure 1-1.  
The ac input to each output board is rectified and applied to a regulator circuit. Each output board employs series regulation  
techniques. A regulator element is connected in series with the load and operates in the linear region (between saturation  
and cutoff) of the transistor characteristic curve. Regulation is achieved by varying the conduction of the series element in  
response to a change in line voltage or circuit load.  
The output board receives digital signals from the GP-IB board and converts them to analog signals which program the  
output voltage, current, and overvoltage values. The output may be programmed remotely over the GP-IB using commands  
(see Chapter 5) or locally from the supplys front panel using the control keys (see Chapter 6).  
The output board can be commanded to send measurement and status data back over the GP-IB and/or front panel. The data  
is sent back via the supplys GP-IB board. GP-IB readback capabilities include output voltage and current, present and  
accumulated status, and all programmed settings. The front panel LCD display can indicate the output voltage and current,  
the supplys GP-IB address, error messages, and programmed values. Annunciators on the front panel indicate the operating  
status of the selected output (output board).  
Specifications  
Table l-2 lists the performance specifications for the Agilent 662xA power supplies. Performance specifications describe  
the instrument’s warranted performance. The service manual, Option 9l0, contains procedures for verifying the  
performance specifications.  
Table 1-3 lists the supplemental characteristics for the Agilent 662xA supplies. Supplemental characteristics are type-tested  
or typical values, which are based on a product sample and, while representative, are not guaranteed.  
Qualifying Conditions  
All performance specifications apply over the full operating temperature range of the power supply (0 to 55°C) unless  
otherwise specified. All regulation, accuracy, etc. specifications are plus or minus the values listed. All measurements are  
made at the rear terminals of the supply with a resistive load and local sensing unless otherwise specified. Voltage  
measurements are made from the + S to the - S terminals. Overvoltage measurements are made from the + V to the - V  
terminals. + Current refers to the output acting as a current source while - Current refers to the output acting as a current  
sink.  
Definitions  
Load effect: Maximum steady state change in the regulated output parameter due to a change in load resistance on the  
output in question.  
Source effect: Maximum steady state change in the regulated output parameter due to a change in the source voltage  
within rated values.  
Cross regulation: Maximum steady state change in the regulated output parameter due to a change in load resistance on  
any other output(s).  
Programming accuracy: (Calibration temp ± 5°C) Maximum difference between the programmed value and the actual  
output.  
Readback accuracy: (Calibration temp ±5°C) Maximum error in reading back an output parameter. Expressed as a  
percentage of the reading, plus a constant.  
General Information 15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Output response time: Beginning at the time the power supply has finished processing a VSET command (change  
output voltage), the maximum time for the output voltage to settle to within a settling band about the final value from any  
specified operating point. This value must be added to the command processing time to obtain total programming time (see  
Figure 1-3). Time constant is the maximum time required for the voltage to reach 63% of its final value.  
Figure 1-3. Output Response Characteristics  
Temperature coefficient: Maximum change in the regulated output parameter per °C change in ambient temperature  
after a 30 minute warmup. Expressed as a percentage plus a constant per degrees C. (Plus a constant for readback  
temperature coefficient).  
Long Term Drift: Maximum change of regulated output voltage or current during an 8-hour period following a 30 minute  
warmup, with all influence and control quantities maintained constant. Expressed as a percentage of the setting plus a  
constant.  
Short Term Drift: Maximum change of regulated output voltage or current within 30 minutes after a line and/or load  
change. Expressed as a percentage of the setting plus a constant.  
Output Noise (PARD): PARD replaces the former term ripple and noise. PARD is the periodic and random deviation of  
dc output voltage or current from its average value, over a specified bandwidth and with all influence and control quantities  
maintained constant.  
Programming resolution: Average programming step size.  
Current Sinking (- Current): Each output can sink as well as source current. The sinking capability is not programmable  
and depends upon the output voltage. The current sinking capability is described in greater detail in  
Chapter 4.  
Table 1-2. Specifications  
PERFORMANCE SPECIFICATIONS: (0 to 55 °C unless otherwise specified)  
DC Output Ranges: All outputs will accept voltage programming commands 1% higher than those listed and  
current programming commands 3% higher than those listed.  
Output Power  
Output Range  
Output Volts  
Output Amps  
25 Watt Output  
Hi Range  
50 Watt Output  
Hi Range  
Lo Range  
0-7 V  
0-15 mA  
Lo Range  
0-16 V  
0-200 mA  
0-50 V  
0-500 mA  
0-50 V  
0-2 A  
Load Effect (Regulation):  
Voltage  
0.5 mV  
0.5 mV  
0.5 mV  
0.5 mV  
+ Current  
0.005 mA  
0.005 mA  
0.01 mA  
0.015 mA  
16 General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 1-2. Specifications (continued)  
Source Effect:  
Voltage  
0.5 mV  
0.5 mV  
0.5 mV  
0.5 mV  
+ Current  
0.005 mA  
0.005 mA  
0.01 mA  
0.01 mA  
Programming Accuracy:  
Note: The programming accuracy specifications may degrade slightly when the unit is subjected to an RF field equal  
to or greater than 3 volts/meter.  
Voltage  
+ Current  
OVP  
0.016% + 1.5 mV  
0.04% + 15 µA  
0.13% + 475 mV  
0.016% + 10 mV  
0.04% + 100 µA  
0.13% + 475 mV  
0.016% + 3 mV  
0.04% + 185 µA  
0.13% + 475 mV  
0.016% + 10 mV  
0.04% + 500 µA  
0.13% + 475 mV  
Measurement/Readback Accuracy*  
Voltage  
0.016% + 2 mV  
0.016% + 10 mV  
0.016% + 3.5 mV  
0.016% + 10 mV  
± Current  
0.03% + 15 µA  
0.03% + 130 µA  
0.04% + 250 µA  
0.04% + 550 µA  
Load Transient Recovery Time :(all outputs)  
75 µS maximum to recover to within 75 mV of nominal value following a load change from 0.1 A to 100% of the  
maximum rated current  
Maximum Output Noise: (PARD) (20 Hz-20 MHz)  
CV pk-to-pk  
CV rms  
+CC rms  
3 mV  
0.5 mV  
0.1 mA  
3 mV  
0.5 mV  
0.1 mA  
3 mV  
0.5 mV  
0.1 mA  
3 mV  
0.5 mV  
0.1 mA  
AC Input Voltage and Frequency:  
Nominal Line l00, 120, 220, or 240 Vac  
Amplitude  
Frequency  
+6% to -13% of nominal line voltage  
47 to 66 Hz  
Note at low line, the supply will operate with up to 3/4 ohm line resistance  
*For a ± 5 ° C range about the calibration temperature  
Table 1-3. Supplemental Characteristics  
25 Watt Output  
Output Power  
Output Range  
Output Volts  
Output Amps  
50 Watt Output  
Lo Range  
0-7 V  
0-15 mA  
Hi Range  
0-50 V  
Lo Range  
0-16 V  
Hi Range  
0-50 V  
0-2 A  
0-500 mA  
0-200 mA  
Temperature Coefficient-Programming:  
Voltage  
+ Current  
OVP  
(0.003%+0.1 mV)  
per° C  
(0.003%+0.1 mV)  
per° C  
(0.003%+0.1 mV)  
per° C  
(0.003%+0.1 mV)  
per° C  
(0.0035% + 1.5 µA)  
(0.0035% + 6 µA)  
(0.0035% + 30 µA)  
(0.0035% + 50 µA)  
per° C  
per° C  
per° C  
per° C  
(0.013% + 1 mV)  
per° C  
(0.013% + 2 mV)  
per° C  
(0.013% + 1 mV)  
per° C  
(0.013% + 2 mV)  
per° C  
General Information 17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 1-3. Supplemental Characteristics (continued)  
Output Power  
Output Range  
Output Volts  
Output Amps  
25 Watt Output  
Hi Range  
50 Watt Output  
Lo Range  
0-7 V  
0-15 mA  
Lo Range  
0-16 V  
0-200 mA  
Hi Range  
0-50 V  
0-2 A  
0-50 V  
0-500 mA  
Temperature Coefficient-Measurement:  
Voltage  
(0.002% + 0.1 mV)  
per° C +0.5 mV  
(0.0025% + 1.5 µA)  
per° C + 1 µA  
(0.002% + 0.1 mV)  
per° C +4 mV  
(0.0025% + 10 µA)  
per° C + 40 µA  
(0.002% + 0.1 mV)  
per° C +1 mV  
(0.0025% + 20 µA)  
per° C + 15 µA  
(0.002% + 0.1 mV)  
per° C +4 mV  
(0.0025% + 50 µA)  
per° C + 150 µA  
± Current  
Long Term Drift: (In an 8 hour period following a 30 minute warm up)  
Voltage  
0.006% + 0.5 mV  
0.006% + 2 mV  
0.006% + 0.5 mV  
0.006% + 2 mV  
+ Current  
0.01% + 5 µA  
0.01% + 20 µA  
0.01% + 40 µA  
0.01% + 60 µA  
Short Term Drift: (Within 30 minutes after a line and/or load change):  
Voltage  
0.002 + 0.5 mV  
0.002 + 1 mV  
0.002 + 0.5 mV  
0.002 + 1 mV  
+ Current  
0.01% + 2 µA  
0.01% + 20 µA  
0.01% + 25 µA  
0.01% + 60 µA  
DC Floating Voltage:  
No output terminal may be more that 240 Vdc from any other terminal or from chassis ground. Also, no overvoltage  
terminal may be more than 240 Vdc from any other terminal or chassis ground.  
Remote Sense Capability: (see pages 50 - 52):  
Outputs can maintain specifications with up to 10 volt total in the remote sense leads.  
AC Input Power  
: 550 Watts  
Programming Resolution:  
Voltage  
+ Current  
OVP  
3.2 mV  
33 µA  
230 mV  
1 mV  
13 µA  
230 mV  
3.2 mV  
131 µA  
230 mV  
460 µV  
1 µA  
230 mV  
Readback Resolution:  
Voltage  
+ Current  
- Current  
3.3 mV  
48 µA  
37 µA  
1.1 mV  
14 µA  
14 µA  
3.3 mV  
160 µA  
151 µA  
483 µV  
1 µA  
1 µA  
Fixed Overvoltage Protection: (Measured at output terminals +V and -V)  
Minimum  
Nominal  
Maximum  
56 V  
60 V  
64 V  
56 V  
60 V  
64 V  
56 V  
60 V  
64 V  
56 V  
60 V  
64 V  
18 General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 1-3. Supplemental Characteristics (continued)  
High Line Inrush Current:  
Peak Value  
100 V Opt  
85 A  
120 V Opt  
85 A  
220 V Opt  
50 A  
240 V Opt  
50 A  
rms Value  
6.3A  
5.7 A  
3.0 A  
3.0 A  
Fuse Rating  
8 A  
8 A  
4 A  
4 A  
GP-IB Interface Capabilities:  
SH1. AH1, T6, L4, SR1, RL1, PP1, DC1, DT0, C0, E1  
Current Sink Capability:  
25 Watt output: 0.50 A  
50 Watt output: 1.0 A (2.0 A below 16 V)  
Command Processing Time: (see Figure 1-3)  
7 milliseconds typical (with front panel disabled). Using STD and RCL commands allows you to change all the  
voltage and current settings in about 10 mS (with front panel disabled).  
Maximum Output Programming:  
Response Time  
Settling Band  
6 mS  
50 mV  
6 mS  
50 mV  
6 mS  
50 mV  
6 mS  
50 mV  
Maximum Time  
Constant  
750 µS  
750 µS  
750 µS  
750 µS  
Series and Parallel operation:  
Two outputs can be operated directly in parallel or can be connected for straight series operation. Refer to Chapter 4  
for more information.  
Reactive Load Capability:  
All outputs have been designed with the ability to operate with significant reactive loads without instability (refer to  
Figures 1-4 through 1-6)  
Safety Agency Compliance:  
This series of power supplies is designed to comply with the following standards:  
IEC 348, UL 1244, and CSA 22.2 No. 231.  
Dimensions (all models):  
Height  
Width  
Depth  
132.6mm (5.22 in)  
425.5mm (16.75 in.)  
497.8mm (19.6 in.)  
General Information 19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 1-3. Supplemental Characteristics (continued)  
Weight:  
Agilent 6625A, 6628A  
Agilent 6626A, 6629A  
Net  
Shipping  
15.5 kg (34 lbs.)  
20.8 kg (46 lbs.)  
17.7 kg (39 lbs.)  
23 kg (51 lbs.)  
Load Cross Regulation:  
Voltage  
+ Current  
0.25 mV  
0.005 mA  
0.25 mV  
0.005 mA  
0.25 mV  
0.01 mA  
0.25 mV  
0.01 mA  
20 General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 1-4. CV Operation with Capacitive Load, Stability Graph for all Outputs  
General Information 21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 1-5. Output Noise (Typical) CC Mode  
Figure 1-6. Output Noise (Typical) CV Mode  
22 General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 1-7. Output Impedance (Typical) Graph (See Supplemental Characteristics)  
General Information 23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2
Installation  
Introduction  
This chapter contains instructions for checking and mounting your power supply, connecting your supply to ac power,  
converting it from one line voltage to another, and connecting the GP-IB cable.  
The power supply generates operating magnetic fields which may affect the operation of other instruments. If your  
instrument is susceptible to magnetic fields, do not locate it in the immediate vicinity of the power supply. Typically, at  
three inches from the power supply, the electromagnetic field is less than 5 gauss.  
Initial Inspection  
Your instrument was thoroughly inspected and tested before it left the factory. As soon as you receive it, remove the power  
supply from its packing case and check to make sure it has not been damaged in shipment. Check that there are no broken  
connectors or keys, and that the cabinet and panel surfaces are free from dents and scratches. Check the rear panel terminal  
blocks and front panel display for any cracks. If damage is found, you should file a claim with the carrier immediately and  
notify the Agilent Technologies Sales and Service office nearest you.  
Chapter 3 of this manual includes an electrical turn-on check-out procedure which, when carried out successfully, will give  
you a high level of confidence that the power supply is operating in accordance with its specifications. Detailed electrical  
checks complete with verification procedures are included in the Service Manual.  
Keep the original packing materials for the carriers inspection if there was damage, or in case any equipment has to be  
returned to Agilent Technologies. Warranty information is printed on the inside cover of this manual. Remember to send a  
detailed description of the failure and symptoms when returning the power supply for service. Your Agilent Technologies  
Sales and Service office will furnish the address of the nearest service office to which the instrument can be shipped.  
Location and Cooling  
Your power supply can operate without loss of performance within the temperature range of 0 to 55 ° C (measured at the  
fan intake). The fan, located at the rear of the unit, cools the supply by drawing air in through the openings on the rear panel  
and exhausting it through openings on the sides. Using Agilent Technologies rack mount kits will not impede the flow of  
air.  
Because the power supply is fan cooled, it must be installed in a location that allows sufficient space at the rear and the  
sides for adequate circulation of air. Either side may be restricted to have as little as 1 inch (25 mm) space.  
Figure 2-1 gives the dimensions of the power supply cabinet. These dimensions apply to both models. The cabinet has  
plastic feet that are shaped to ensure self-alignment when stacked with other Agilent Technologies System II cabinets. The  
feet may be removed for rack mounting.  
The power supply can be mounted in a standard 19 inch rack panel or enclosure. Rack mounting accessories for this unit  
are listed under options (page 11) of Chapter 1. Complete installation instructions are included with each rack mounting kit.  
Instrument support rails are required for non-stationary installations. These are normally supplied with the cabinet and are  
not included with the rack mounting kits.  
Installation 25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 2-1. Outline Diagram  
Input Power Requirements  
You can operate this power supply from a nominal 100 V, 120 V, 220 V or 240 V single phase power source at 47 to 66  
Hz. The input voltage range, maximum input current, high line inrush current (PK), and the fuse required for each of the  
nominal inputs are listed in Table 2-1. You can check the line voltage setting of your supply by examining the door on the  
line module. This is located on the rear panel of your supply as shown in Figure 2-2. The red mark that appears in one of  
the four windows on the line module indicates the line voltage setting for which your supply is set.  
If necessary, you can convert the supply from one line voltage setting to another by following the instructions under LINE  
VOLTAGE CONVERSION (see page 29).  
Table 2-1. Input Power  
Nominal  
Voltage  
100 V  
120 V  
220 V  
Line Voltage Range  
Maximum Input Current  
High Line Inrush  
Current (PK)  
85 A  
Fuse  
(rms.)  
6.3 A  
5.7 A  
3.0 A  
3.0 A  
8 AM  
8 AM  
4 AM  
4 AM  
Nominal  
-13%, +6%  
85 A  
50 A  
50 A  
240 V  
Line Fuse  
The ac line fuse is located behind the door on the line module (see Figure 2-3). To access the fuse, remove the power cord  
and push against the tab on the line module in the direction of the ac input socket. The current rating of the fuse is based on  
the line voltage setting of your supply. Table 2-2 gives the Agilent part numbers for the fuses that should be used with  
specific line voltages.  
26 Installation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 2-2. Rear Panel Detail (Agilent 6626A Shown)  
Table 2-2 Line Fuses  
Line  
Fuse  
Needed  
8AM  
Agilent Part Number (for ¼ X 1¼ in. fuses  
Voltage  
100/120 V  
220/240 V  
only)  
2110-0342  
2110-0055  
4AM  
Note: All fuses are rated for 250 V.  
Figure 2-3. Line Module Detail  
Installation 27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Power Cord  
The power supply is shipped from the factory with a power cord that has a plug appropriate for your location. Figure 2-4  
shows the standard configuration of plugs used by Agilent Technologies. Below each drawing is the Agilent part number  
for the replacement power cord equipped with a plug of that configuration. If a different power cord is required, contact the  
nearest Agilent Technologies Sales and Service office.  
For your protection, the National Electrical Manufacturers Association (NEMA) recommends that the instrument panel and  
cabinet be grounded. This power supply is equipped with a three-conductor power cord; the third conductor being the  
ground. The power supply is grounded only when the power cord is plugged into an appropriate receptacle. Do not operate  
this power supply without adequate cabinet ground connection.  
Figure 2-4. Power Cord Plug Configurations  
SHOCK HAZARD. Connect the power cord to a grounded receptacle before you connect any  
external floating voltages to the supply.  
The offset pin on the standard three-prong power cord connector is the ground connection. If a two contact receptacle is  
encountered, it must be replaced with a properly grounded three-contact receptacle in accordance with the National  
Electrical Code, local codes and ordinances. The work should be done by a qualified electrician.  
Line Voltage Conversion  
You can change the supply to accept 100 V, 120 V, 220 V and 240 V ac input by adjusting the voltage selector card located  
inside of the line module (see Figure 2-3). After you have changed the line voltage, refer to Table 2-2 and check that the  
fuse inside the line module is the correct fuse for that line voltage. The procedure is as follows:  
1. Turn off power and remove the power cord from the ac input socket on the back of the power supply.  
2. To open the line module, move the plastic door on the module aside. If your line voltage change requires a change in the  
rating of the fuse, rotate FUSE PULL to the left and remove the fuse.  
3. Grasp the voltage select pc board with a pair of needle-nose pliers and slide it out of its slot.  
4. To select a voltage, orient the pc board so that the desired voltage appears on the top left side of the board. Push the  
board all the way back into its slot. The desired line voltage must be visible when the board is installed.  
5. Install the correct fuse in the door of the line module if your line voltage change also requires a change in the rating of  
the fuse (see Table 2-2).  
28 Installation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
FIRE HAZARD. Make sure the replacement fuse is one of the same type (size) and rating (amps) that  
is consistent with the voltage level you are operating at. Do not use a substitute fuse; use a fuse with  
the same Agilent Part number listed in Table 2-2.  
6. Close the door of the line module and insert the power cord in the ac input socket. Your power supply is now configured  
to operate at the voltage you selected.  
GP-IB Interface Connector  
The GP-IB connector on the rear panel connects your power supply to your computer and other GP-IB devices (see Figure  
2-2). Accessories (page 12) in Chapter 1 lists the cables and cable accessories that are available from Agilent Technologies.  
An GP-IB system can be connected together in any configuration (star, linear, or both) as long as the following rules are  
observed  
1. The total number of devices, including the computer, is no more than 15.  
2. The total length of all the cables used is no more than two meters times the number of devices connected together, up to a  
maximum of 20 meters.  
Note  
IEEE Std. 488-1978 states that you should exercise caution if your individual cable lengths exceed 4 m.  
Do not stack more than three connector blocks together on any GP-IB connector. The resultant leverage can exert excessive  
force on the mounting panels. Make sure that all connectors are fully seated and that the lock screws are firmly finger  
tightened. Do not use a screwdriver. Use a screwdriver only for the removal of the screws.  
Installation 29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3
Getting Started  
Introduction  
This chapter is intended for the first time user of the supply. It provides four main discussions:  
Front Panel Controls and Indicators  
Turning on Your Supply  
Checking Out Your Supply Using Local Control  
Introduction to Remote Operation  
First, the supplys front panel controls and indicators are briefly described. Some of the controls and indicators will be used  
in the Turn On And Checkout procedures that follow. Chapter 6 describes how to use all of the front panel controls.  
Successful completion of the Turn On And Checkout procedures ensures with a high level of confidence that your supply is  
operating properly. Complete performance testing and troubleshooting procedures are given in the Service Manual (Agilent  
Part No. 06626-90003).  
The checkout procedures are performed locally from the front panel. In addition to checking the operation of your supply,  
these simple step-by-step checkout procedures will help the first time user become familiar with operating the supply from  
the front panel.  
When you have completed the checkout procedures, you are then introduced to the fundamentals of operating the supply  
remotely from a computer. You will learn how to send a command to the supply from the computer and how to get data  
back to the computer from the power supply. A few of the most often used power supply commands will be described to  
help you get started and become familiar with the basics of programming your supply.  
After completing this chapter, you can proceed to Chapter 4 to find out how to make load connections to your supplys  
outputs and then to Chapter 5 (Remote Control) and/or Chapter 6 (Local Control) to learn all the details about operating  
your supply.  
Front Panel Controls and Indicators  
The power supplys controls and indicators are shown in Figure 3-1 and are described in Table 3-1. Note that the front panel  
controls are identical for all models except for the number of OUTPUT annunciators (number 3 in Figure 3-1).  
Getting Started 31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3
5
1
6
7
8
6626A SYSTEM DC POWER SUPPLY  
SYSTEM  
ADDR  
ERR  
OUTPUT  
ENTRY  
VOLTS  
AMPS  
LCL  
OUTPUT  
SELECT  
RANGE  
V/I  
7
4
1
0
8
5
2
.
9
6
3
VOLT  
VOLT  
45.153 14.235m  
OVSET  
VSET  
ISET  
1
2
3
4
CV CC UNR OCP ERR RMT ADDR SRQ  
ENBLD  
-- OUTPUT --  
STO  
OCP  
ENTER  
CURR  
CURR  
OUTPUT  
ON/OFF  
RCL  
RESET  
LINE  
ON  
OFF  
9
4
2
Figure 3-1. Agilent 6626A Front Panel  
Table 3-1. Controls and Indicators  
Description  
Number  
Controls/Indicators  
Page  
1
Returns power supply to local mode (unless local  
3-13, 5-2,  
6-1  
LCL key  
lockout has been received via GP-IB). Also, turns the  
power supply’s display on if it was turned off via the  
GP-IB.  
2
RMT - Indicates that the power supply is operating  
under remote control (GP-IB)  
3-13, 5-2,  
6-1,  
GP-IB Status  
Annunciators  
(These three  
annunciators indicate the ADDR - Indicates that the power supply is addressed to  
3-10, 5-1  
GP-IB status of the  
power supply).  
talk or to listen.  
SRQ - Indicates that the power supply is requesting  
service.  
5-1, 5-2,  
5-3, 5-17  
3
4
Indicate which output channel has been selected for  
front panel control and/or display (Only one output  
annunciator can be on at a time.)  
3-7, 3-8,  
6-1--6-4  
OUTPUT  
Annunciators  
CV - Indicates that the selected output channel is in the  
constant voltage mode.  
3-8, 4-1,  
6-2, 6-3  
Power Supply Status  
Annunciators  
(These five annunciators CC - Indicates that the selected output channel is in the  
3-8, 4-1,  
6-2, 6-3  
indicate the status of the  
power supply).  
positive constant current mode ( + CC) or the negative  
current limit ( - CC) mode.  
32 Getting Started  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 3-1. Controls and Indicators (continued)  
Number  
4 (cont)  
Controls/Indicators  
Description  
UNR- Indicates that the selected output channel is  
unregulated; i.e., it is not regulated by CV or CC control  
loops.  
Page  
4-4  
OCP ENBLD - Indicates that the overcurrent protection 3-8, 6-4,  
function for the selected channel is enabled.  
5-13  
ERR - Indicates that a programming or hardware error  
has occurred and that the ERR bit in the serial poll  
register has not been cleared  
4-4, 5-16  
Normally displays the measured output voltage and  
Display (When power is current for the selected channel. When programmed  
3-7--3-10,  
4-4, 5-19,  
5-20,  
5
6
Alphanumeric LCD  
turned on, all segments  
will be displayed for  
approximately 2  
seconds).  
from the front panel, the function being programmed  
(e.g. VSET), the output channel (e.g. 2), and the present 6-1--6-4  
value (e.g. 2.250) will be displayed. Error conditions  
will be spelled out in alpha characters  
ADDR - Displays the power supplys GP-IB address.  
(These four control keys You can change the address using the numeric entry  
3-10, 5-2,  
6-4  
System Control Keys  
affect the entire power  
supply and are  
keys. You cannot query or change the address remotely  
(over the GP-IB).  
independent of the  
output selected.)  
ERR - Displays a programming or hardware error  
message and clears the ERR bit in the serial poll register  
5-1, 5-20  
5-1, 5-13  
STO - Used in conjunction with the numeric entry keys  
to store the present output conditions for all outputs in  
the specified internal register (0-10). Each register stores  
the following parameters: VSET, VRSET, ISET,  
IRSET, and OVSET. In addition, register 0 also stores  
OCP, DLY, and MASK parameters. Register 0-3 are  
non-volatile.  
The parameters stored in register 0 are the "power-up"  
parameters. These parameters are set upon applying AC  
power to the power supply. Note that the STO  
command cannot be used over the bus more than once  
per non-volatile register (0-3) without cycling the AC  
power (OFF and ON). However, these registers can be  
used for an unlimited number of times via the front  
panel keys without cycling the AC power .  
RCL - Used in conjunction with the numeric entry keys 5-1, 5-13  
to recall the settings from the specified internal register  
(0 to 10). All outputs are set to the recalled values.  
Getting Started 33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 3-1. Controls and Indicators (continued)  
Number  
Controls/Indicators  
Description  
Page  
7
OUTPUT SELECT - Selects one of the output  
3-7, 3-8,  
Output Control Keys  
(These twelve keys are  
output dependent).  
channels for local control or display. This key allows the 6-1--6-4  
channels to be selected in forward (Ð) or reverse (Ï)  
sequence.  
VSET - Displays the selected outputs present voltage  
setting. The setting can be changed using the numeric  
entry keys, Ñ VOLT, or Ò VOLT keys..  
3-8, 3-9,  
5-9, 5-11,  
6-2  
ISET - Displays the selected outputs present current  
setting. The setting can be changed using the numeric  
entry keys, Ñ CURR, or Ò CURR keys.  
3-8, 5-11,  
6-3  
OUTPUT ON/OFF - Toggles the selected output on  
and off. When off, "DISABLEDappears on the  
display.  
5-12, 6-3  
OVSET - Displays the selected outputs overvoltage trip 3-8, 5-12,  
point. The setting can be changed using the numeric  
entry keys.  
6-4  
RANGE V/I - Displays the selected outputs full scale  
programming range for voltage and current. This key  
also enables the Up/Down keys to change the  
programming range.  
3-9  
Up/Down Keys - These keys increment or decrement  
the voltage or current settings of the output selected.  
The display will be in the metering mode for the output  
selected. When one of the keys is pressed, it will change  
the output selected by one least significant bit (LSB). If  
the key is pressed and held, it will continually change  
the setting until it is released The Ñ VOLT key is  
disabled when the supply is in the CC mode. The Ò  
CURRENT key is disabled when the supply is in the CV  
mode.  
3-9, 6-2,  
6-3  
These keys will also allow you to change the  
programmed resolution range of voltage or current after  
the RANGE V/I key is depressed.  
3-9  
Ñ VOLT - Increases the selected output voltage by an  
LSB and then a faster rate as the key is kept pressed, or  
after the RANGE V/I key has been pressed, sets the  
selected output to the high voltage range.  
3-9, 6-2  
Ò VOLT - Decreases the selected output voltage by an  
LSB and then at a faster rate as the key is kept pressed,  
or after the RANGE V/I key has been pressed, sets the  
selected output to the low voltage range.  
3-9, 6-2  
34 Getting Started  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 3-1. Controls and Indicators (continued)  
Controls/Indicators Description  
Number  
Page  
7 (cont)  
Ñ CURRENT - Increases the selected output current by 3-9, 6-3  
an LSB and then at a faster rate as the key is kept  
pressed, or after the RANGE V/I key has been pressed,  
sets the selected output to the high current range.  
Ò CURRENT - Decreases the selected output current  
by an LSB and then at a faster rate as the key is kept  
pressed, or after the RANGE V/I key has been pressed,  
sets the selected output to the low current range.  
3-9, 6-3  
OCP - Toggles the selected outputs overcurrent  
protection circuit on and off.  
3-8  
3-8  
RESET - This key performs the following:  
1. Reset the selected outputs overvoltage crowbar (the  
cause of the overvoltage must be removed before reset is  
successful) .  
2. Reset the selected outputs overcurrent condition and  
returns the output to its previous settings (the cause of  
the overcurrent must be removed before reset is  
successful).  
3. Return the display to the metering mode from any  
other mode (e.g. VSET). In the metering mode, the  
measured output voltage and current of the selected  
output are displayed .  
8
0 to 9 - Set the value of the specified function and  
3-8,  
6-1--6-5  
Numeric Entry Keys  
(These keys are used in  
conjunction with many  
of the System Control  
and Output Control keys  
to enter the desired  
and  
(e.g. VSET 1 6.550  
.
Ï(backspace) - Erases the previous keystroke.  
Depressing this key without setting a  
value places the display in  
6-1  
values into the power the  
metering mode. Supply.  
ENTER - Enters the values on the display for the  
specified function, initiates the function, and returns the  
display to the metering mode. Pressing this key without  
setting a value will result in retention of the previous  
values and returning the display to the metering mode.  
3-8,  
6-1--6-5  
9
Turns ac power on and off.  
3-6  
LINE switch  
Table 3-1, in addition to providing a brief description of each control and indicator, lists the paragraphs in which the use of  
each control and indicator is described. Because most of the functions performed by the front panel controls can also be  
performed remotely by power supply commands, the corresponding paragraphs in Chapter 5 (Remote Operation) are listed  
in Table 3-1 where applicable.  
Getting Started 35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Turning On Your Supply  
The following paragraphs describe the power-on sequence which includes a self test of most of the power supplys circuits.  
Before you turn on your supply, make sure that:  
The line module on the rear panel is set to match your input line voltage.  
The proper fuse is installed and the line cord is plugged in.  
If you have any questions concerning installation or power requirements, review Chapter 2.  
To turn on your supply, press the front panel LINE switch. When the power is initially applied, the supply performs a series  
of self tests which last about 3 seconds. Included in these tests are checks of circuits on the GP-IB board and on each of the  
output boards.  
Normal Self Test Indications  
If the supply passes the self test, the display will first show all segments of the LCD display with annunciators on as  
illustrated in Figure 3-2.  
Figure 3-2. Test Pattern of all Display Segments at Power-on  
After all segments are displayed, the supplys GP-IB address will appear for approximately 2 seconds as shown in Figure 3-  
3. As shipped from the factory, the power supplys address is set to 5. You must know this address before you can remotely  
program your supply.  
Figure 3-3. Typical Address Display During Power-On Self Test  
When self test is successfully completed, the power-up parameters stored in register 0 will be set (see page 89 for a  
complete description of the store and recall functions). As shipped from the factory, this non-volatile register contains zero  
volts and zero current (see Table 5-4 on page 72) for all outputs. Note that the CV annunciator will indicate that the output  
is in the constant voltage mode.  
36 Getting Started  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 3-4. Typical Display at Power-On  
Self-Test Errors  
If the supply fails the power-on self-test, all power supply outputs will remain disabled (off) and the display will indicate  
the type of failure and the output channel on which it occurred. Figure 3-5 shows that self-test detected an error in output  
channel 3. Error messages that could appear on the display if self-test fails are listed below. Self-test error messages are  
explained in Appendix D and troubleshooting procedures are given in the Service Manual. You may also call your Agilent  
Sales Office for help.  
Power-On Self Test Error Messages  
HDW ERR CH "N"  
8291 FAILED  
TIMER FAILED  
CV DAC CH N"  
CC DAC CH "N"  
OV DAC CH "N"  
FUSE CH N"  
Note: "N" specifies the failed output channel number: 1,2,3, or 4 as applicable.  
Figure 3-5. Sample Self-Test Failure Display  
Checking Out Your Supply Using Local Control  
The following procedures use the display and keys on the front panel to check each of your power supplys outputs. No test  
equipment, other than a jumper wire (14 AWG), is required to perform these tests. The tests must be repeated for each  
output of your particular supply. The checkout consists of voltage, overvoltage, and current tests. It is assumed that power  
has already been turned on, the supply has passed the power-on self-test, loads are not connected to any of the supplys  
outputs, and sense clips are connected between the sense terminals and the output terminals.  
Note  
The following procedures are identical for all models and for all outputs. Use the OUTPUT SELECT key  
to select an output to be tested. If an output fails any of the tests, refer to the troubleshooting section in  
the Service Manual.  
Getting Started 37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Voltage Test  
1. Set the voltage of the selected output to 10 V by pressing:  
VSET  
1
0
ENTER  
2. Check that the display reads approximately 10 V and 0 A and the CV annunciator is on indicating that the supply is in  
the constant voltage mode of operation.  
Overvoltage Test  
1. Program the overvoltage protection (OVP) to 19 V by pressing:  
OV  
SET  
1
9
ENTER  
ENTER  
2. Set the voltage to 16 V by pressing:  
VSET  
1
6
3. Check that the display reads approximately 16 V and 0 A.  
4. Set the voltage to 20 V by pressing:  
VSET  
2
0
ENTER  
5. Check that the display reads "OVERVOLTAGE’  
6. Reset the supply by pressing:  
VSET  
1
6
ENTER  
RESET  
7. Check that the display reads approximately 16 V and 0 A.  
Current Test  
1. Turn off the supply.  
2. Remove the barrier block cover from the output to be tested and connect a short circuit (jumper wire) between the + V  
and -V terminals of the output being tested.  
3. Turn on the supply.  
4. Use the  
to select the output being tested.  
5. Set the voltage to a volts by pressing:  
VSET  
5
ENTER  
0
Set the current to 0.01 amps by pressing:  
ISET  
.
1
ENTER  
38 Getting Started  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Check that the display reads approximately 0 volts and the programmable current limit value. Also, check that the front  
panel CC annunciator is on indicating that the output is in the constant current mode of operation.  
7. Set the current to 0.5 A by pressing:  
ISET  
.
5
ENTER  
8. Check that the display reads approximately 0 V and 0.5 A.  
9. Enable the overcurrent protection circuit by pressing:  
OCP  
10. Check that the OCP ENBLD annunciator is on indicating that overcurrent protection is enabled and the display reads  
’’OVERCURRENT’. When in overcurrent, the output is disabled.  
11. Disable the overcurrent protection circuit by pressing:  
OCP  
12. Reset the output by pressing:  
RESET  
13. Check that the display reads approximately 0 V and 0.5 A.  
14. Turn off the supply and remove the jumper from the output terminals.  
Repeat the voltage, overvoltage and current tests given on page 38 for the other output channel(s) using the  
key  
Changing Resolution Range  
1. Press the  
RANGE V/I  
key.  
2. Observe the selected outputs full scale programming range for voltage and current in the front panel display.  
3. Press either the Ñ or the Ò key. (voltage or current)  
4. Observe that the selected outputs full scale programming range for voltage and current changes to the high range (if the  
Ñ key is pressed) or low range (if the Ò key is pressed). Note that no range change will occur if the Ñ or Ò key is pressed  
while the supply is already in the requested range.  
Introduction To Remote Operation  
The following paragraphs explain the fundamentals of operating the supply remotely from a computer. Only a few  
commonly used programming commands will be discussed. Refer to Chapter 5 for a detailed description of all the  
commands. The intent of this discussion is to help first time users to quickly become familiar with operating their supply  
from a computer.  
The programming examples that follow assume that a computer is connected to the GP-IB connector on the rear of your  
supply (see Chapter 2), power is applied, and loads are not connected to any of the supplys outputs. The examples used are  
primarily for Agilent Series 200/300 computers using Agilent BASIC language. Read the manuals for your particular  
computer to find out which statements you must use.  
Getting Started 39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Enter/Output Statements  
The programming statements you use to operate your supply from remote depend on your computer and its language. In  
particular, you need to know the statements your computer uses to output and enter information. For example, the Agilent  
BASIC language statement that addresses the power supply to listen and sends the command to the power supply is:  
OUTPUT  
The Agilent BASIC language statement that addresses the power supply to talk and reads back data from the power supply  
is:  
ENTER  
The supplys front panel ADDR annunciator is on when the supply is addressed to talk or to listen.  
Reading the GP-IB Address  
Before you can operate your power supply remotely, you need to know its GP-IB address. The address was displayed  
during the power on sequence described in page 36. To see the address, press:  
ADDR  
A typical address display is shown in Figure 3-6:  
Figure 3-6. Typical Address Display  
The displayed response is the power supplys GP-IB address. When sending a remote command, you append this address to  
the computers GP-IB interface select code (normally 7). For example, if the select code is 7 and the power supplys GP-IB  
address is 5, the combination is 705.  
Changing the GP-IB Address  
Note  
All examples in this discussion assume a GP-IB address of 5. It is recommended that you retain this  
address to simplify programming.  
Every device on the GP-IB must have an address. The supplys address is factory set to decimal 5. Any address from 0  
through 30 is a valid address. If you need to change the Agilent 662xA Supplys address press:  
ADDR  
You can now enter a new address. For example, press:  
1
4
ENTER  
You have now changed the address from 5 to 14. If you want to change the address back to 5, repeat the above procedure  
but use 5 instead of 14 in the last step. Note that the address is stored in the power supplys non-volatile memory and  
therefore will be retained through interruption of the ac line power.  
40 Getting Started  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Sending a Remote Command  
To send the power supply a remote command, combine your computers output statement with the GP-IB interface select  
code, the GP-IB device address, and finally, the power supply command. For example, to set the output voltage of output  
channel 1 to 2 volts, send:  
Getting Data From The Supply  
The supply is capable of measuring the values of its output parameters in response to queries. In this example, the query  
asks the supply to measure the output voltage at output 1.  
When you send a query from remote, the supply does not display the response as it did when you executed the command  
from the front panel. Instead, it holds the response in an output buffer. The output buffer is a register that holds information  
until it is read by the computer or is replaced with new information.  
Note  
On an Agilent Series 200/300 Computer, the A variable must be declared before you do the following  
steps. Refer to your computers operating manual for more information.  
Use your computers enter statement to get the response from the output buffer. For example, execute:  
ENTER 705; A  
Followed by:  
DISP A  
The ENTER statement enters whatever is in the supplys output buffer into the computers A variable. The DISP statement  
displays the A variables contents on the computers display.  
Often Used Commands  
The command set contains over sixty commands that allow you to program the power supply in a variety of applications.  
Within this command set, however, is a small subset of commands that are all you need for most applications. These  
commands are: VRSET, IRSET, VSET, ISET, VOUT?, IOUT?, OUT, OVSET, and OCP.  
Getting Started 41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Each of these commands is briefly discussed in the following paragraphs to help you get started in programming your  
supply. To know more about these commands, refer to Chapter 5.  
The VRSET and IRSET commands select the range the power supply operates in. Two ranges are available for each output  
- standard resolution, and high resolution (see Table 1-2).  
If the value for voltage or current entered is not within the present operating range of the supply, the error annunciator will  
indicate an error. Sending ERR?, and addressing the supply to talk, will return the error message (in the case number  
range" error would be displayed.  
Voltage and Current Programming. You can send voltage and current values to the power supply directly in volts or  
amps. The following examples use voltage and current values that are within the range of any output that the power supply  
provides.  
To set the voltage of output 1 to 5 volts, send:  
OUTPUT 705; “VSET 1,5”  
To set the current of output 2 to 450 milliamps, send:  
OUTPUT 705; ''ISET 2,.450"  
Output Voltage and Current Measurement. The power supply has built-in metering circuitry, which allows it to monitor  
the output voltage and current of the selected output. You can instruct the supply to return its present output voltage and  
current using VOUT? and IOUT? queries respectfully. The metering circuitry is auto-ranging, and will select the proper  
resolution.  
To measure the output voltage at output 1, send:  
OUTPUT 705; "VOUT? 1"  
To get the measurement from the output buffer, send:  
ENTER 705; A  
DISP A  
The computer should display a reading of approximately 5 volts .  
To measure the output current at output 2, send:  
OUTPUT 705: "IOUT? 2''  
To get the measurement from the output buffer, send:  
ENTER 705; A  
DISP A  
42 Getting Started  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Output On/Off. You can turn a specified output on or off. Individual outputs can be controlled as shown below.  
To turn off output 1, send:  
OUTPUT 705; OUT 1,0"  
When an output is turned off, it is set to 0 volts and to the minimum current limit value  
To turn on output 1, send:  
OUTPUT 705; OUT 1,1"  
When an output is turned on it will return to the voltage and current settings determined by the present VSET and ISET  
values.  
Overvoltage Setting. You can send an overvoltage setting value to the power supply directly in volts. If the output voltage  
exceeds this setting, the output crowbar is fired, and the output voltage is quickly downprogrammed and disabled (0 volts  
output).  
To set the overvoltage value of output 2 to 3.5 volts, send:  
OUTPUT 705; “OVSET 2,3.5”  
Overcurrent Protection. The output will go to the off state (0 volts and 0 current) when the overcurrent protection (OCP)  
feature is enabled and the output is in the + CC mode. To enable the overcurrent protection mode for output 2 send:  
OUTPUT 705; ''OCP 2,1"  
To disable the overcurrent protection mode for output 2, send:  
OUTPUT 705; ''OCP 2,0"  
When overcurrent protection is disabled and the output is in + CC mode, the output current will be limited to and will stay  
at the ISET value.  
Returning the Supply to Local Mode  
In the remote mode (RMT annunciator on), the front panel keys have no effect on any of the supply's outputs and only the  
computer can control the supply. However, you can still use the front panel display to monitor the output voltage and  
current or check any of the present settings (VSET, ISET, OVSET, etc.) of the selected output channel.  
If you want to use the front panel keys to change the output settings, you must return the supply to the local mode. You can  
return the supply to the local mode (provided that the local lockout command has not been received from the computer) by  
pressing the LCL key. A change between the local and remote modes (or vice versa) will not result in a change in the power  
supply outputs. Refer to Chapter 6 for additional details on using the LCL key and operating the supply in the local mode.  
Getting Started 43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4
Output Connections and Operating Information  
Introduction  
This chapter explains how to make connections to the output terminals located on the rear of your power supply. Some  
general operating information is included in this chapter to help you understand how the power supply operates under  
various load conditions. This information applies whether you are operating the supply via the front panel or the GP-IB.  
Output Ranges  
Figure 4-1 identifies the output combinations that are available on the power supply. Each output can operate as a constant  
voltage (CV) or constant current (CC) source over a wide variety of output voltage and current combinations. In addition,  
each output has an active downprogrammer which operates at currents up to approximately 110% of the maximum positive  
current rating of the output. This means that each output can actively sink as well as source its maximum rated output  
current. At voltages below 2.5 V, a downprogramming resistor continues downprogramming until the voltage reaches  
approximately zero volts.  
Figure 4-1. Output Combinations  
Output Connections and Operating Information 45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operating Quadrants  
Figures 4-2A and 4-2B show the operating locus of your power supply in three quadrants. The area in quadrant 1 shows the  
operating locus defined by the voltage and current settings of each output. The area in quadrant 2 indicates the locus where  
each output can operate as a current sink. You cannot program current limit values in quadrant 2.  
Figure 4-2. Typical Output Range Characteristics  
46 Output Connections and Operating Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 4-3 shows the current sink characteristics lower voltages in greater detail. The area in quadrant 4 illustrates the  
reverse polarity diode characteristics of each output. Do not operate any output with reverse-voltage currents that are  
greater than the maximum rating of the output.  
Figure 4-3. Typical Downprogramming Characteristic Below 2.0 V  
Operating Ranges  
Notice in Figure 4-2, there are four overlapping areas in quadrant one for both the 25 watt and 50 watt outputs. The Low  
Range Voltage and Low Range Current ranges are higher precision than the High Range Voltage and High Range Current  
ranges. The voltage and current ranges can be set independently (see page 73). If the high resolution range is programmed,  
attempting to program values outside this range will result in a number range error.  
The 50 watt output (Figure 4-2B) is different from the 25 watt output (Figure 4-2A), in that there is a power limiting  
boundary at 1 amp in the High Voltage/High Current range. As a result, the 50 watt output can be set to no more than 1.03  
amps at voltage settings above 16.16 volts, and vice versa.  
Setting parameters of voltage or current beyond this power limit boundary, but within the High Voltage/High Current range  
will cause the other parameter (voltage or current) to automatically be scaled down to the power limiting boundary (see  
pages 71 and 73).  
The voltage and current programming resolution of the outputs is affected by the setting of the range. The voltage of the 25  
watt output, when programmed to the Low Voltage Range, has a voltage programming resolution of 460 µV. When  
programmed above 7 volts, the High Voltage Range must be selected resulting in a programming resolution of 3.2 mV.  
The current of the 25 watt output, when programmed to the High Current Range, has a current programming resolution of 1  
µA. When programming above 15 mA, the High Current Range must be selected, resulting in a programming resolution of  
33 µA.  
The voltage of the 50 watt output, when programmed to the High Voltage Range, has a voltage programming resolution of  
1 mV. When programming above 16.16 volts, the high range must be selected, resulting in a programming resolution of 3.2  
mV.  
The current of the 50 watt output, when programmed to the High Current Range, has a current programming resolution of  
13 µA. When programmed above 200 mA, the High Current Range must be selected, resulting in a programming resolution  
of 131 µA.  
The readback range of the outputs is not programmable, and is selected automatically by the power supply.  
Output Connections and Operating Information 47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The readback resolution of the 25 watt outputs when metering voltages of 7 volts or below, will be 483 µV. For voltages  
above 7 volts, the readback resolution will be 3.3 mV.  
The readback resolution of the 25 watt outputs when metering source currents of 15 mA or below, the readback resolution  
will be 1 µA. When metering source currents above 15 mA, the readback resolution will be 48 µA.  
When metering sink currents of 15 mA or below, the readback resolution of the 25 watt outputs will be 1 µA. When  
metering sink currents above 15 mA, the readback resolution will be 37 µA.  
When metering voltages of 16 volts, or below, the readback resolution of the 50 watt outputs will be 1.1 µV. For voltages  
above 16 volts. the readback resolution will be 3.3 mV.  
When metering source currents of 200 mA or below, the readback resolution of the 50 watt outputs will be 14 µA. When  
metering source currents above 200 mA, the readback resolution will be 160 µA.  
When metering sink currents of 200 mA or below, the readback resolution of the 50 watt outputs will be 14 µA. When  
metering sink currents above 200 mA, the readback resolution will be 151 µA.  
Protection Features  
Protective circuitry within the supply can limit or turn off an output in the event of an abnormal condition. The activated  
protection feature can be determined by observing the front panel display area. You can also read back the status of the  
supply over the GP-IB. The following protection features are implemented:  
OVERVOLTAGE -- shorts the output by firing an SCR crowbar and sets zero volts and minimal current on an output if  
any of the following conditions are present:  
1. The output voltage exceeds the programmed overvoltage trip point.  
or  
2. The sum of the voltage from the + V output terminal to the +S terminal plus the voltage from the -S terminal to the -V  
output terminal exceeds 10.0 V (applies to remote sensing only).  
or  
3. A trip signal is received on the outputs OV terminals.  
or  
4. The outputs fixed overvoltage circuit is activated.  
The OV trip point can be programmed up to 55 V. When an overvoltage occurs, the word OVERVOLTAGE appears in the  
front panel display and the OV status bit is set for that output. Chapter 5 explains how to program the overvoltage trip level.  
A fixed overvoltage threshold of approximately 120% of the maximum rated output voltage is built into each output.  
Because the fixed overvoltage circuit is biased from the output terminals, it can be activated and provide protection even  
when the supply is not connected to the ac power line.  
The OVRST command or front panel RESET key restores the programmed voltage and current values and clears the OV  
once the cause of the overvoltage has been eliminated.  
OVERCURRENT -- when the overcurrent protection feature is enabled, and the output is sourcing current and enters the +  
CC operating mode, the output will be disabled (set to zero volts and minimal current) and the word OVERCURRENT will  
appear on the front panel display. In addition, the OC status bit is set for that output. The OCRST command or front panel  
RESET key restores the programmed voltage and current values and clears the OC once the cause of the overcurrent  
condition has been eliminated. Refer to Chapter 5 for programming details.  
48 Output Connections and Operating Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
UNREGULATED OUTPUT -- the supply informs the user when output regulation is not guaranteed. This can occur when  
attempting to sink excessive currents below 4 volts on 25 W outputs and 2 volts on 50 W outputs or when operating outputs  
in parallel. The UNR annunciator on the front panel and the UNR bit in the status register indicate that the specified output  
is unregulated. Line voltage dropout or an incorrectly set ac power module can also cause the output to become  
unregulated. If line voltage dropout continues, the supply shuts down and will return to the power-up condition when  
normal line voltage is restored.  
OVERTEMPERATURE -- shuts down the linear pass transistors and downprogrammer of the output that has reached an  
unsafe operating temperature. Operation of the other outputs is unaffected. An overtemperature can occur because of  
excessively high ambient temperature, a blocked fan, or insufficient space at the sides for adequate air circulation. When an  
overtemperature condition occurs, the word OVERTEMP appears in the front panel display and the OT status bit is set.  
This circuit resets automatically and restores the output approximately 30 seconds after the temperature drops sufficiently  
for safe operation.  
ERROR -- if the power supply receives an invalid command either through the front panel or the GP-IB, the ERR  
annunciator on the front panel comes on and the ERR bit in the serial poll register is set. The power supply does not execute  
the command and remains at previously set values. Pushing the ERR button in local mode displays the error message and  
clears the error. The error indicator may also indicate that an instrument failure has occurred. Refer to Appendix D for  
further details.  
Connecting The Load  
Each terminal block cover on the rear panel is secured by a locking tab which snaps into a slot at the left of the terminal  
block. To remove, insert a screwdriver into this rectangular slot and move the locking tab to the left. When the locking tab  
releases, gently pull the terminal block cover away from the terminal block. To reinstall the cover, align it over the terminal  
block and gently press it into position until the locking tab engages.  
SHOCK HAZARD. Turn off ac power before making rear panel connections. All wires and straps  
must be properly connected with terminal block screws securely tightened. Replace terminal block  
covers before reapplying power.  
Each rear terminal block has six M3.5 x 0.6 x 6 mm screws for attaching wires (see Figure 2-2). Load connections to the  
supply are made at the + V and -V terminals on each terminal block. Do not connect unterminated wires to the load  
terminals. Wires used for load connections must be properly terminated with termination connectors securely attached.  
Remember to replace the impact resistant plastic covers (Agilent P/N 06624-20007) over the terminal blocks after making  
connections.  
Consistent with good engineering practice, all sense and trigger leads connected to the rear terminal blocks should be  
twisted and shielded to maintain the instruments specified performance.  
Wire Size Selection  
FIRE HAZARD Select a wire size large enough to carry short-circuit current without overheating.  
Two factors must be considered when selecting wire size for load connections: conductor temperature  
and voltage drop. To satisfy safety requirements, load wires must be heavy enough not to overheat  
while carrying the short-circuit output current of the unit. Table 4-1 lists the current-carrying capacity  
(ampacity) for various sizes of stranded wire.  
Note that the minimum wire size required to prevent overheating may not be large enough to prevent OV trip when remote  
sensing and to maintain good regulation. The load wires should be large enough to limit the total voltage drop in both leads  
to no more than 10 volts. See Figure 4-6 for total allowed load lead drop when programming to high voltage settings. With  
remote sensing, load regulation is degraded 1 mV per 1 V in the - V output terminal load lead. (See Remote Sense  
Connections on page 52).  
Output Connections and Operating Information 49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Note  
To prevent tripping of the overvoltage circuit, pick a wire size sufficient to handle the FULL output  
current of the unit no matter what the intended load current or current limit setting.  
Table 4-1 lists the resistivity for various wire sizes and the maximum lengths to limit the voltage drop to 1.0 volts for  
various currents.  
Table 4-1. Stranded Copper Wire Ampacity and Maximum Wire Lengths to Limit Load Lead Voltage Drop  
Ampacity Per Wire (Amps)  
Max Length to Limit Voltage  
Drop to 1 V Per Lead  
Wire Size  
2 Wire Bundled  
4 Wire Bundled  
Resistivity  
5 A  
10 A  
(feet)  
10  
15  
25  
20 A  
(AWG)  
(/ft)  
20  
18  
16  
14  
7.8  
6.9  
0.0102  
0.0064  
0.0040  
0.0025  
0.0016  
20  
30  
50  
--  
5
7.5  
12.5  
20  
14.5  
18.2  
29.3  
37.6  
12.8  
16.1  
25.9  
33.2  
40  
12  
--  
--  
30  
(Cross Section  
Area in mm2)  
(/m)  
(meters)  
0.5  
0.75  
1
1.5  
2.5  
7.8  
9.4  
12.7  
15.0  
23.5  
6.9  
8.3  
11.2  
13.3  
20.8  
0.0401  
0.0267  
0.0200  
0.0137  
0.0082  
5
7.4  
10  
14.6  
--  
2.4  
3.8  
5
7.2  
12.2  
1.2  
1.8  
2.6  
3.6  
6
Notes:  
1. Ampacities for AWG wires are derived from MIL-W-5088B. Maximum ambient temp: 55°C. Maximum wire temp:  
105°C.  
2. Ampacities for metric wires are derived from IE Publication 335-1.  
3. Ampacity of aluminum wire is approximately 84% of that listed for copper wire.  
4. Because of wire inductance considerations, it is recommended that you keep your load leads twisted, tie wrapped, or  
bundled together and less than 50 feet (14.7 meters) in length per lead.  
5. See Page 49 for information on wire gauge considerations with capacitive loads.  
Note  
The OVP circuit senses at the main output terminals and not on the sense leads. Thus, the voltage sensed  
by the OVP circuit could be as much as 10 V higher than the voltage being regulated at the load. Program  
the OVP trip voltage accordingly when using remote sensing. In addition, if the total voltage drop in both  
leads exceeds 10 V, a protective circuit will fire the OVP circuit regardless of the OVP setting.  
Load lead resistance is an important factor relating to the CV stability of the supply with remote sensing of capacitive  
loads. If high capacitance loads are expected, you should not use wire gauges heavier than 12 to 14 AWG for long runs of  
load lead. See Figure 1-4 for more information about stability with output capacitors.  
Multiple Loads  
If you will be sensing the load locally (as shipped from the factory) and are connecting multiple loads to one output,  
connect each load to the output terminals using separate connecting wires (see Figure 4-4). This minimizes mutual coupling  
effects and takes full advantage of the power supply's low output impedance. Each pair of wires should be as short as  
possible and twisted or bundled to reduce lead inductance and noise pickup.  
If load considerations require the use of distribution terminals that are located remotely from the supply, connect the power  
supply output terminals to the remote distribution terminals by a pair of twisted or bundled wires. Connect each load to the  
distribution terminals separately. Remote voltage sensing is recommended under these circumstances. Sense either at the  
remote distribution terminals or, if one load is more sensitive than the others, directly at the critical load.  
50 Output Connections and Operating Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 4-4. Optimum Hookup for Multiple Loads, Local Sensing  
Note  
When a load is connected through relay or switch contacts, contact bounce may activate the overvoltage  
circuit and shut down the supply. Therefore, it is recommended that the output be downprogrammed to 0  
or turned-off (disabled) before the relay (or switch) contact is opened or closed.  
Positive and Negative Voltages  
Either positive or negative voltages can be obtained from the supply by grounding (or "commoning") one of the output  
terminals. Always use two wires to connect the load to the supply regardless of where or how the system is grounded. This  
supply can be operated with any output terminal ± 240 Vdc (including output voltage) from ground.  
Remote Voltage Sensing  
The power supplies will allow up to a total load lead drop of 10 V.  
Because of the unavoidable voltage drop developed in the load leads, the as-shipped terminal block strapping pattern shown  
in Figure 4-4 does not provide the best possible voltage regulation at the load. The remote sensing connections shown in  
Figure 4-5 improve the voltage regulation at the load by monitoring the voltage there instead of at the supplys outputs  
terminals. This allows the power supply to automatically compensate for the voltage drop in the load lead. Remote sensing  
is especially useful for CV operation with load impedances that vary or have significant lead resistance. It has no effect  
during CC operation. Because sensing is independent of other power supply functions, remote sensing can be used  
regardless of how the power supply is programmed. Note that with remote sensing, voltage readback monitors the load  
voltage at the sense points.  
Output Connections and Operating Information 51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 4-5. Remote Voltage Sensing  
The maximum voltage available at the power supply output terminals during remote sensing (see Figure 4-6) is 50.5 volts.  
This allows the sum of the voltage across both load leads to equal 10 volts maximum. For lower output voltages refer to  
Figure 4-3.  
Remote Sense Connections  
Remember to turn off the power supply before making or changing any connections on the rear panel terminal blocks.  
Connect the unit for remote sensing by first disconnecting the straps between sense and load terminals. Then make your  
connections as shown in Figure 4-5. Connect the sense leads as close to the load as possible. See wire size selection  
(pages 49 - 50) for information on selection of load lead wire gauge. Best results will be obtained by using the shortest load  
leads practical. It is recommended that you keep your load leads under 14.7 meters (50 feet) per lead because of inductance  
effects.  
52 Output Connections and Operating Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 4-6. Total Allowable Load Lead Voltage Drop (total of both leads) with Remote Sensing  
Output Noise Considerations  
Any noise picked up on the sense leads will appear at the supplys output and may adversely affect CV load regulation.  
Twist the sense leads or use a ribbon cable to minimize the pickup of external noise. In extremely noisy environments it  
may be necessary to shield the sense leads. Ground the shield at the power supply end only; do not use the shield as one of  
the sensing conductors.  
The noise specifications in Table 1-2 apply at the power supply output terminals when using local sensing. However,  
voltage transients may be produced at the load by noise induced in the leads or by load current transients acting on the  
inductance and resistance of the load lead. If it is desirable to keep voltage transient levels to a minimum, place an  
aluminum or a tantalum capacitor, with an approximate value of 10 µF per foot (30.5 cm) of load lead, right across the load  
(see Figure 4-5). Refer to Figure 1-4 for capacitive load stability considerations.  
Programming Response Time with an Output Capacitor  
Because voltage programming into an external output capacitor may cause the supply to briefly enter CC operating mode,  
voltage programming response time may be longer than that specified in Table 1-2. Use the following formula to estimate  
the additional response time:  
(Added Output Capacitor)(Charge in Vout)  
Additional Response Time =  
Current Limit Setting  
Output Connections and Operating Information 53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Open Sense Leads  
The sense leads are part of the supplys feedback path. Connect them in such a way so that they do not inadvertently  
become open circuited. The power supply includes protection resistors that reduce the effect of open sense leads during  
remote-sensing operation. If the sense leads open during operation, the supply returns to the local sensing mode, with the  
voltage at the output terminals approximately equal to the programmed value.  
Overvoltage Trigger Connections  
Each output of your power supply has two OV terminals on its rear panel terminal block. These terminals are labeled +OV  
and -OV. By connecting the OV terminals all in parallel as shown in Figure 4-7, an overvoltage shutdown on any one  
output will also trigger the overvoltage on the remaining outputs. Any number of OV terminals up to eight sets can be  
strapped together. Observe polarity when connecting the OV terminals in parallel.  
Figure 4-7 Overvoltage Connections  
The overvoltage trip point for each output can be set either from the front panel or by remote programming. You can also  
externally fire the overvoltage circuit of one or more outputs by applying a 5 volt pulse of at least 50µs to any pair of OV  
terminals (see Figure 4-8). As long as all OV terminals are wired together, the outputs will be crowbarred simultaneously .  
External Trigger Circuit  
Figure 4-8 illustrates a recommended external circuit that can be used to provide an OV trip signal to the OV terminals.  
This circuit configuration provides good noise immunity and protects against the voltage pulse that is returned from the OV  
terminals every time that the overvoltage circuit fires. It can be operated from a wide range of bias voltages provided the  
input limiting resistors are chosen as tabulated in the figure. If it is not required to trip the OV with a TTL signal, then a  
bias supply, switch, current limiting resistor (R2), and protection diode are all that are required. Note that with the unit off  
(ac power removed), the + OV and - OV terminals are inactive.  
The internal equivalent OV circuit is shown in Figure 4-9. Note the internal DC blocking capacitor, bleed resistor, and  
noise bypass capacitors.  
Do not exceed 50 volts maximum between the + OV and the - OV terminals. The OV terminals are  
rated at ±240 Vdc (including external OV voltage) from chassis ground or any other output terminals.  
54 Output Connections and Operating Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 4-8. External Trigger Circuit  
Figure 4-9. Equivalent Internal OV Trigger Circuit  
Power Supply Protection Considerations  
Battery Charging  
If you are using your supply in a battery charging application, it is recommended that a series protection diode be added to  
prevent damage to the supply during an overvoltage shutdown. Remember that each output has an overvoltage protection  
circuit that fires a crowbar to disable the output for any of the OVERVOLTAGE conditions described on page 48.  
Figure 4-10 illustrates the recommended connections and protection circuit for a battery charging application. The diode  
will prevent damage to your supply that can result from excessive battery current flowing into the supplys output in the  
event of an overvoltage shutdown.  
Output Connections and Operating Information 55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 4-10. Recommended Protection Circuit for Battery Charging  
Capacitive Load Limitation  
The programmable overvoltage protection circuit can be used to downprogram capacitive loads although it is primarily  
intended for use as a protection feature as described on page 48.  
Repetitive (over 100 cycles) tripping of the overvoltage circuit with output capacitors greater than 2000  
µF may result in eventual damage to the supply.  
Parallel Operation  
Connect in parallel only outputs that have equivalent voltage and current ratings.  
Connecting outputs in parallel provides a greater current capability than can be obtained from a single output. Because each  
output contains an active downprogrammer that is capable of sinking current from only ONE identical output, you can  
parallel no more than two outputs. These outputs must have equivalent voltage and current capability. For example, on the  
Agilent 6626A power supply, only outputs 1 and 2 or 3 and 4 may be connected together. On the Agilent 6625A supply,  
none of the outputs may be connected together. If you are mixing different model supplies (i.e. an Agilent 6625A and an  
Agilent 6626A) be sure to connect only equivalent voltage and current outputs in parallel.  
As an example, Figure 4-11 shows how to connect two outputs in parallel to a single load with local sensing. This  
configuration applies to both CV and CC operating modes. Connecting the load leads of output 2 directly to the + V and - V  
terminals of output 1 keeps the total length of the load leads to a minimum and reduces the number of wire connections that  
must be made at the load itself. Connecting the + S and -S terminals of output 2 directly to the sense terminals of output 1  
compensates for the IR drop in the interconnecting load leads.  
56 Output Connections and Operating Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 4-11. Parallel Connections with Local Sensing  
CV Operation  
For CV operation, one output must operate in CC mode and the other output must operate in CV mode. Although each  
output operates independently of the other, the output that is operating in CV mode will be controlling" the voltage  
regulation of both outputs. Setting the output voltages as outlined in the following paragraph and configuring the outputs as  
shown in Figure 4-11 will allow output 1 to operate in CV mode and output 2 to operate in CC mode.  
To assure that output 2 will be operating in CC mode, you must program output 2s voltage to a higher value than the  
voltage of output 1. One way to accomplish this is to first program output 2 to the maximum allowable voltage setting for  
the desired operating range (see Table 4-2 or Figure 4-2). These values are 1% higher than the rated voltage for the  
operating range. Then, program output 1s voltage to the desired operating voltage. The lower voltage setting of output 1  
will determine the voltage that appears across the load. The current limit point of the paralleled outputs is the sum of both  
individual current limit points. The output current of the parallel combination is the algebraic sum of the individual current  
readbacks.  
The + OV and - OV terminals of output 1 should be wired to the + OV and - OV terminals of output 2. When programming  
the overvoltage setpoint, set both outputs to the same overvoltage value. When resetting the overvoltage, first disable both  
outputs by using the OUTPUT ON/OFF key or OUT command. Next, reset both overvoltages. Finally, re-enable the  
outputs with the OUTPUT ON/OFF key or OUT command.  
Table 4-2. Maximum Allowable Voltage Setting  
Output Type  
25W  
Maximum Low Range Voltage  
Maximum High Range Voltage  
7.07V  
16.16V  
50.5V  
50.5V  
50W  
Note  
Below 4 V on 25 W outputs and 2 V on 50 W outputs, the downprogrammer cannot sink the maximum  
rated current. (See Figures 4-2 and 4-3). To operate parallel outputs at voltages under 2.5 V, program  
both outputs to the same voltage setting. Depending on the load, one output may operate in the  
unregulated mode.  
Output Connections and Operating Information 57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CC Operation  
For CC operation, set the output voltages as outlined in CV operation (page 57), or alternatively, program the voltage  
settings of both outputs to the same voltage limit point. Then program the current of each output so that the sum of both  
currents equals the total desired operating current. The simplest way to accomplish this is to program each output to one  
half of the total desired operating current. Both outputs will operate in the CC mode.  
Remote Sensing  
If it is necessary to remote voltage sense at the load, parallel the sense leads of output 1 with the sense leads of output 2 and  
connect to the load as shown in Figure 4-12. The outputs can be programmed as previously described.  
Figure 4-12. Parallel Connections with Remote Sensing  
Specifications for Parallel Operation  
Specifications for outputs operating in parallel can be obtained from the specifications for single outputs. Most  
specifications are expressed as a constant or as a percentage (or ppm) plus a constant. For parallel operation, the percentage  
portion remains unchanged while constant portions or any constants are changed as indicated below.  
Current  
All parallel specifications referring to current are twice the single output  
specification except for programming resolution which is the same for both  
single output and parallel output operation.  
Voltage  
All parallel specifications referring to voltage are the same as for a single  
output except for CV load effect, CV load cross regulation, CV source effect,  
and CV short term drift. Below 4 V on 25 W outputs and 2 V on 50 W  
outputs, these are all twice the voltage programming accuracy (including the  
percentage portion). CV load effect above 4 V on 25 W outputs and 2 V on  
50 W outputs, could be twice the load effect specification for a single output.  
CV output noise for output voltages less than 4 V on 25 W outputs, and 2 V  
on 50 W outputs, may be slightly higher than the output noise for a single  
output.  
Load Transient Recovery  
Time  
350 µs maximum to recover within 100 mV of nominal value following a load change  
within the range of 0 to full load.  
58 Output Connections and Operating Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Series Operation  
SHOCK HAZARD. Floating voltages must not exceed 240 Vdc. No output terminal may be more  
than 240 Vdc from chassis ground.  
Connect in series only outputs that have equivalent current ratings. Each output has a reverse voltage  
protection diode across its output terminals. The current conducted by this diode is not internally  
limited by the output. Therefore, never connect an output in such a way that this  
diode will conduct  
current in excess of the rated current of the output since damage could result.  
Connecting outputs in series provides a greater voltage capability than can be obtained from a single output. Because the  
current is the same through each element in a series circuit, outputs connected in series must have equivalent current  
ratings. Otherwise, the higher rated output could potentially damage the lower rated output by forcing excessive current  
through it under certain load conditions.  
Figure 4-13 shows an example of how to connect two outputs in series to a single load with local sensing. This  
configuration applies to both CV and CC operating modes. Connecting the + load lead of output 2 directly to the  
- V terminal of output 1 completes the series connection between the two outputs. Connecting the + S terminal of output 2  
directly to the - S terminal of output 1 and removing the sense jumper (between + S and + V) on output 2 compensates for  
the IR drop in the load lead from output 2 to output 1.  
Figure 4-13. Series Connections with Local Sensing  
CV Operation  
For CV operation, first program the current setting of each output to the desired current limit point. Then program the  
voltage of each output so that the sum of both voltages equals the total desired operating voltage. The simplest way to  
accomplish this is to program each output to one half of the total desired operating voltage. Both outputs will operate in CV  
mode.  
Output Connections and Operating Information 59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CC Operation  
For CC operation, the current setting of each output must be programmed to the desired operating current. The sum of the  
voltage settings determines the voltage limit point. As an example, one way to program the voltage of the output is to set  
the voltage of each output to one half of the total voltage limit point. Then, at load voltages less than one half of the total  
voltage limit point, one output will operate in CC mode while the other output will be conducting through its internal  
reverse voltage protection diode. At load voltages greater than one half the total voltage limit point, the output that was in  
CC mode will change to CV mode while the output that was conducting through its diode will regulate the current in CC  
mode and provide the balance of the voltage required by the load. Note that the total load voltage can be found by adding  
the results of reading back the individual series outputs only when neither reverse voltage protection diode is conducting.  
When this diode is conducting, the corresponding output has reverse voltage across it so that its voltage readback may not  
be accurate.  
When an output is conducting through its reverse voltage protection diode, the output will have a reverse voltage across its  
output terminals with the - V terminal more positive than the + V terminal. This voltage will be maximum at the rated  
current of the output. (See Figure 4-2 for reverse diode characteristic). Note that when an output conducts through this  
diode, it will indicate CC mode even though it is not regulating the current or voltage. Also, note that the voltage readback  
is not specified to indicate negative voltages although it will operate down to a limit of about - 0.176 V on the 25 W outputs  
and -0.392 V on the 50 W outputs. These values will still be indicated even if the actual voltage is more negative.  
Remote Sensing  
If it is necessary to remote voltage sense at the load, connect the sense leads of output 1 and output 2 as shown in Figure 4-  
14. Note that the + sense lead of output 2 must remain connected to the -sense terminal of output 1. The outputs may be set  
as previously described.  
Figure 4-14 Series Connections with Remote Sensing  
Specifications for Series Operation  
Specifications for outputs operating in series can be obtained from the specifications for single outputs. Most specifications  
are expressed as a constant or a percentage (or ppm) plus a constant. For series operation, the percentage portion remains  
unchanged while constant portions or any constants are changed as indicated below.  
60 Output Connections and Operating Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Voltage  
Current  
All series specifications referring to voltage are twice the single output specification  
except for programming resolution which is the same as for a single output.  
All series specifications referring to current are the same as for a single output except  
for CC load effect, CC load cross regulation, CC source effect, and CC short term drift  
which are twice the current programming accuracy (including the percentage portion).  
Load Transient  
Recovery Time  
Load transient recovery time is the same to within approximately twice the voltage  
setting band since the output impedances of the series combination add together.  
Bi-Polar Operation  
Figure 4-15 shows an example connecting the outputs of the power supply for bi-polar operation to test power FETs  
(PFET).  
Great flexibility is achieved with this setup. The outputs of the power supply can be programmed as voltage or current  
sources providing positive or negative voltages or currents. By taking measurements via internal readback circuitry in the  
power supply, the outputs can be used as voltmeters or ammeters.  
In order to properly measure the parameters of the PFET, it is desirable to be able to directly regulate any two of the bias  
parameters of the PFET, while being able to measure a third. The setup shown in Figure 4-15 allows control of Vds and Ids  
while measuring Vgs.  
In this example, outputs 3 and 4 are connected to allow bipolar operation. Notice that the - S and - V leads, of outputs 3 and  
4, are strapped together. The + S lead of V3 senses the voltage at the drain of the Device Under Test (DUT) and the + S  
lead of output 4 senses the voltage at the source of the DUT. Outputs 1 and 2 are strapped to sense the gate to source  
voltage.  
Figure 4-15. Bipolar Operation  
Output Connections and Operating Information 61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5
Remote Operation  
Introduction  
Chapter 3 introduced you to the basics of remote operation and provided a few simple examples using a Series 200/300  
computer as the GP-IB controller. This chapter contains all the information required to control your power supply remotely  
and discusses in greater detail how each of the commands can be implemented. The material covered is intended for any  
controller capable of using the GP-IB interface functions mentioned in Interface Functions on page 63.  
Four major sub-sections are discussed. These are:  
1. GP-IB Operation  
2. Programming Syntax  
3. Initial Conditions  
4. Power Supply Commands  
The GP-IB section briefly describes the GP-IB interface functions to get you acquainted with remote programming using  
the GP-IB. Under Programming Syntax, the syntax of all commands, the numeric data formats and the programmable  
ranges for all models are given. Initial Conditions highlights the initial values of all the parameters at power-on. Power  
Supply Commands will describe all the commands which can be used to program the supplys functions including status  
reporting, error handling, protection features, and voltage and current programming. The power supply commands are  
summarized in Appendix C.  
The examples are written in a generic form to make adaptation to your controller easy. You should be familiar with the use  
of your controller and its instruction set and how the power supply commands can be incorporated in your controller  
commands. If you are not familiar with the operation of the power supply, you are advised to read through Chapters 2  
through 4 first.  
GP-IB Operation  
Interface Functions  
Remote control is implemented by the GP-IB. It enables instructions to be sent from an external computer equipped with a  
GP-IB interface. The power supply implements the following IEEE-488 Interface Functions:  
SH1 - Source Handshake  
AH1- Acceptor Handshake  
T6 - Talker  
SR1 - Service Request  
RL1 - Remote/Local  
RL1 - Parallel Poll  
DC1 - Device Clear  
L4 - Listener  
The source handshake, acceptor handshake, talker and listener functions are implemented by the interface circuits of the  
power supply and the controller. The ADDR annunciator indicates when the power supply is addressed to listen or talk.  
(The talker function includes the Serial Poll, see page 64).  
Service Request. This is a message which can be initiated by the power supply to request service from the controller.  
When the supply is requesting service, it asserts the service request (SRQ) line on the GP-IB to interrupt the controller  
providing the controller is configured to service interrupts. A service request can be generated whenever there is a fault on  
one of the outputs (up to 4 outputs), a programming error has occurred, or at power on providing certain commands are  
sent. Service request commands are discussed in detail in Service Request Generation on page 79.  
The SRQ annunciator on the front panel display is turned on when the power supply is requesting service from the  
computer and remains on until the controller conducts a serial poll. A serial poll removes the service request and turns off  
Remote Operation 63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
the SRQ annunciator regardless of whether the condition that caused the service request continues to exist. The service  
request is also removed when you send the "CLRcommand (see page 76).  
Remote/Local. The power supply can receive programming information either from the GP-IB (remote) or from the front  
panel (local). When the power supply is in remote, the state of the supply cannot be changed by using the front panel keys,  
although the LCL key will remain enabled. Remote operation takes precedence over local operation, hence if the supply is  
accepting commands remotely and you attempt to change it to local operation, the supply will not allow any local settings  
and will remain in remote. You can prevent the front panel from sending programming information by sending the local  
lockout command. This command is sent only from the GP-IB. If you change from local to remote or vice-versa, there will  
be no change in the programmed settings.  
Parallel Poll. Parallel Poll allows the controller to receive at the same time one bit of data from each of up to eight  
instruments connected to the bus. Agilent power supplies designate bit #6, the RQS bit of the serial poll register for this  
operation. By checking the status of this bit, the computer can quickly determine which instruments on the bus requested  
service. Once an instrument is identified, the computer can perform a serial poll to find out the exact cause of the request.  
Parallel Poll does not reset this service request bit (RQS) in the power supply.  
Note  
IEEE-488 does not define what data an instrument should put on a bus in response to parallel poll. Many  
instruments such as Agilent Technologies power supplies indicate the state of their RQS bit, but the  
operator should not assume that all instruments on the bus respond to parallel poll with their RQS bit.  
Unless remotely configured, the power supply will respond with a 1 on one of the GP-IB data lines if it is requesting  
service and its address is between 0 and 7. Addresses 0 through 7 define which data line (1 through 8) the supply will  
respond on. If the address is set to 8 or greater, the supply will not respond unless remotely configured.  
The power supply may be remotely configured to respond with a 0 or 1, on any of the data lines, to indicate that it is  
requesting service. This is done in accordance with IEEE-488 1978.  
Serial Poll. In a serial poll, the controller polls each instrument on the bus one at time. The power supply responds by  
placing the contents of the eight-bit serial poll register on the GP-IB data lines. The Serial Poll register is discussed on  
page 78 and defines the function of each of the bits. After the serial poll, the service request is cleared and the SRQ  
annunciator at the front panel is reset (off). However, the condition that generated the service request may still be present.  
See page 79.  
Device Clear. The Device clear command is typically used in systems to send all devices in the system to a known state  
with a single command. It may be implemented as an addressed or an unaddressed command. The power supply CLR  
command performs the same function as Device Clear (see page 76).  
GP-IB Address Selection  
You can find out the present address or change the address of the supply by using the front panel ADDR key as described  
in Chapter 3. Any address 0 through 30 is a valid address. If you program an address outside this range you will get a  
number range error.  
Note  
Care should be taken to not select the controller address.  
64 Remote Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Power-On Service Request (PON)  
The power supply can request service from the controller when the power is turned on. This request can be enabled or  
disabled by sending a PON command (see page 80). When the request is enabled, the supply can generate an SRQ at  
power-on or when there is a momentary loss in power. You can execute a serial poll to clear the service request. Table 5-7  
details the conditions under which a PON command will generate an SRQ.  
Note  
The power supply has a non-volatile memory in which it stores certain system variables. Some of these  
variables are the calibration constants, the present supply address, and the present setting of the PON  
command.  
Programming Syntax  
The following paragraphs describe the syntax of the device command that is used to program your power supply. As shown  
in Figure 5-1, the device command is a specific part of the program statement that your computer will accept. The first part  
of the statement is computer as well as programming language specific. Figure 5-1 shows the structure of a typical  
programming statement for an Agilent Series 200/300 computer. If you are using a different computer or programming  
language, refer to your computer programming manual to determine the correct syntax for this portion of the program  
statement. This section of the manual is only concerned with the device command portion (the part inside the quotes for  
Series 200/300 computers with BASIC) of the program statement.  
Figure 5-1. Typical Program Statement for Series 200 Computers  
Table 5-1. Power Supply Commands  
Command  
Header  
Output  
Channel  
1, 2, 3, 4  
Data Range  
(Fig. 5-2)  
---  
Syntax  
Query accumulated status register  
Return power supply to turn on state  
Turn calibration mode on or off  
Query if calibration mode is on  
Set the state of outputs at power on  
ASTS?  
CLR  
Q2  
C1  
C2  
Q1  
---  
---  
---  
CMODE  
CMODE?  
DCPON  
0,1 (OFF,ON)  
See Table 5-4  
1, 2, 3, 4  
---  
0,1 (OFF, ON) CC+  
2,3 (OFF, ON) CC-  
---  
Query outputs power on state  
DCPON?  
DLY  
---  
1, 2, 3, 4  
1, 2, 3, 4  
---  
Set the reprogramming delay time  
Query the setting of the delay time  
Turn front panel display on or off  
Display a string on front panel  
Query if front panel display is on  
0-32  
---  
C4  
Q2  
C2  
C6  
Q1  
DLY?  
DSP  
0,1(OFF,ON)  
’’S T RI N G ’  
---  
DSP  
---  
DSP?  
---  
Remote Operation 65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 5-1. Power Supply Commands (continued)  
Command  
Header  
Output  
Channel  
---  
Data Range  
(Fig. 5-2)  
See Table 5-8  
Syntax  
Query present hardware error  
Query fault register  
ERR?  
FAULT?  
ID?  
Q1  
Q2  
Q1  
C4  
Q2  
C5  
C3  
C3  
Q2  
C5  
C3  
C3  
C3  
C3  
C4  
Q2  
C4  
Q2  
C4  
C2  
Q1  
C5  
C4  
Q2  
C3  
C4  
Q2  
C3  
C3  
C4  
Q2  
C2  
Q1  
C2  
C4  
Q2  
Q1  
Q2  
C2  
Q1  
1, 2, 3, 4  
---  
---  
---  
Query the model number of supply  
Program the I DAC in counts  
Query setting of I DAC in counts  
Send data to calibrate I circuits  
Sets output to high I cal. value  
Sets output to low I cal. value  
Query measured I output  
IDAC  
IDAC?  
IDATA  
IHI  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
---  
See Service Manual  
---  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
1-4  
ILO  
IOUT?  
IRDAT  
IRHI  
Used to Calibrate I readback circuits  
Set output to + I readback high cal value  
Set output to - I readback high cal value  
Set output to - I readback low cal value  
Set output to + I readback low cal value  
Set full scale current range of output  
Query full scale current range  
Set current of an output  
IRHN  
IRLN  
IRLO  
IRSET  
ISET?  
ISET  
Query current of an output  
ISET?  
ISTEP  
METER  
METER?  
NIDAT  
OCP  
Increase or decrease output current by value  
Select which output will be metered  
Query which output is being metered  
Send data to calibrate -I readback  
Enable overcurrent protection  
Query if OCP is enabled  
---  
---  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
---  
See Table 5-4  
0,1 (OFF,ON)  
---  
OCP?  
Reset overcurrent protection  
OCRST  
OUT  
---  
Turn output on or off  
0,1 (OFF,ON)  
---  
Query if output is on or off  
OUT?  
OVCAL  
OVRST  
OVSET  
OVSET?  
PON  
Perform overvoltage calibration  
Reset overvoltage circuit  
See Table 5-4  
---  
Set overvoltage trip value  
See Table 5-4  
See Table 5-4  
0,1 (OFF,ON)  
---  
Query overvoltage trip value  
Enable power on service request  
Query if PON is enabled  
PON?  
RCL  
1, 2, 3, 4  
---  
Recall voltage and current settings  
Set readback DAC to a number of counts  
Query readback DAC count setting  
Query revision date of ROM  
0-10  
RDAC  
RDAC?  
ROM?  
SROM?  
SRQ  
1, 2, 3, 4  
1, 2, 3, 4  
See Service Manual  
---  
---  
Query revision date of secondary ROM  
Set causes for generating a service request  
Query causes which will generate an SRQ  
1, 2, 3, 4  
---  
---  
01,2,3  
SRQ?  
---  
---  
66 Remote Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 5-1. Power Supply Commands (continued)  
Command  
Store present output state  
Header  
Output  
Channel  
Data Range  
(Fig. 5-2)  
0-10  
Syntax  
STO  
STS?  
C2  
Q2  
C1  
C4  
Q2  
C4  
Q2  
C5  
C3  
C3  
C4  
Q2  
C5  
C3  
C3  
C4  
Q2  
Q2  
Q2  
C4  
Query preset status of output  
1, 2, 3, 4  
Perform self test on GP-IB interface  
Set bits in mask register  
TEST?  
UNMASK  
UNMASK?  
VDAC  
VDAC?  
VDATA  
VHI  
.
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
1, 2, 3, 4  
0-255  
Query bits set in mask register  
Program the voltage DAC in counts  
Querys setting of voltage DAC in counts  
Send data to calibrate the voltage circuits  
Set output to high V calibration value  
Set output to low V calibration value  
Query inputs of analog multiplexer  
Query measured value of an output  
Calibrate the voltage readback circuitry  
Set output to V readback high cal value  
Set output to V readback low cal value  
Set full scale voltage programming range  
Query full scale voltage programming range  
Set output voltage  
See Service Manual  
See Table 5-4  
See Table 5-4  
See Table 5-4 V  
1-18  
VLO  
VMUX?  
VOUT?  
VRDAT  
VRHI  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
See Table 5-4  
VRLO  
VRSET  
VRSET?  
VSET  
Query setting of output voltage  
VSET?  
VSTEP  
Increase or decrease output voltage by value  
Figure 5-2 shows the possible syntax forms for the device commands that are used to program the power supply. Syntax  
forms for the calibration commands that are discussed in Appendix A are also included. The oblong shape at the left of the  
syntax forms contains the command header which must be entered as shown in Tables 5-1 and 5-2. Commands are accepted  
in either uppercase or lowercase letters (ASCII characters). Circles contain characters that must be entered exactly as  
shown. Characters such as a space <SP> or a comma are used to separate elements in the command string. Characters such  
as a line feed < LF > or a semicolon are used to terminate the command string. Rectangles contain parameters that follow  
the command header lines and arrows indicate the correct paths through the syntax diagrams.  
Numeric Data  
The power supply will accept numeric data in implicit point, explicit point, or scientific notation. A general syntax diagram  
for numeric data is included in Figure 5-2. Implicit point notation means that numbers do not contain a decimal point;  
integers for example. Numbers written in explicit notation contain a decimal point, such as 12.35. In scientific notation, the  
letter E stands for "10 raised to". For example, 1.2E3 is read as 1.2 times 10 raised to the 3rd power, which equals 1,200.  
Plus and minus signs are considered numeric characters and are optional. If you program a number with an accuracy that is  
greater than the resolution of the supply, the number will automatically be rounded to the nearest multiple of the power  
supplys resolution. Table 5-1 gives the ranges for numeric data that is sent to the supply.  
The power supply will also return numeric data (ASCII characters) to your computer. The format of the numbers returned  
depends upon the type of data requested. Table 5-2 gives the format for data returned to the computer in response to any of  
the queries that are listed.  
Remote Operation 67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 5-2 (Sheet 1 of 2). Syntax Forms For Power Supply Commands  
68 Remote Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure 5-2 (Sheet 2 of 2). Syntax Forms For Power Supply Commands  
Remote Operation 69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 5-2. Power Supply Queries  
Query  
Header  
(Note 7)  
Channel  
(Note 1)  
Response  
(Notes 5 and 6)  
Initial Value  
Syntax  
(Fig.  
5-2)  
Q2  
Voltage Setting  
Current Setting  
Full Scale Current Range  
Full Scale Voltage Range  
Voltage Output  
Current Output  
OVP Setting  
OC Protection On/Off  
Output On/Off  
Unmask Setting  
Delay Setting  
VSET?  
ISET?  
1,2,3,4  
1,2,3,4  
1,2,3,4  
1,2,3,4  
1,2,3,4  
1,2,3,4  
1,2,3,4  
1,2,3,4  
1,2,3,4  
1,2,3,4  
1,2,3,4  
1,2,3,4  
1,2,3,4  
1,2,3,4  
--  
SZD.DDD(Note 3)  
0(Note 8)  
10 mA (Note 8)  
High (Note 8)  
High (Note 8)  
(Note 2)  
(Note 2)  
ZD.DDD (Note 8)  
SZD.DDD (Note 3)  
Q2  
Q2  
Q2  
Q2  
Q2  
Q2  
Q2  
Q2  
Q2  
Q2  
Q2  
Q2  
Q2  
Ql  
Q1  
Q1  
IRSET?  
VRSET?  
VOUT?  
IOUT?  
OVSET?  
OCP?  
OUT?  
UNMASK?  
DLY?  
--  
--  
(Note 2)  
SZZD.DD  
55 V (Note 8)  
ZZD  
ZZD  
ZZD  
--  
--  
0 (Note 8)  
.02 (Note 8)  
<sp>ZD.DDD  
Status  
Accumulated Status  
Fault  
STS?  
ZZD  
ZD  
ZZD  
ZZD  
ZZD  
ZZD  
ZZD  
--  
--  
--  
ASTS?  
FAULT?  
ERR?  
SRQ?  
PON?  
Error  
--  
0 (OFF)  
0 (OFF) (Note 9)  
1 (ON)  
Service Request Setting  
Power-On SRQ On/Off  
Display On/Off  
Model Number  
--  
--  
--  
--  
DSP?  
ID?  
Q1  
Q1  
Agilent 662XA (Note  
--  
4)  
Selftest  
Calibration Mode  
DC Power On  
TEST?  
CMODE?  
DCPON?  
--  
--  
--  
ZZD  
ZZD  
ZZD  
--  
Q1  
Q1  
Q1  
0 (OFF)  
1 (ON) (Note 9)  
S = Sign  
NOTES:  
Z = Digit with leading zeros put out as spaces  
D = Digit  
< sp > = space  
1. Output channels 3 and 4 are not used in all modes. (See Table 5-4).  
2. Current is SZD.DDDDD (0.5 A) & SZD.DDDDDD (15 mA) SZD.DDDD(2 A) & SZD.DDDDD  
(0.2 A)  
3. Voltage can be SZD.DDDD on 7 VR  
4. "X" depends upon model.  
5. A space is returned for a + sign.  
6. All responses are followed by a <CR> and <LF> (EOI asserted with <LF>).  
7. Spaces are allowed between the header and the question mark.  
8. Unit powers up with values in state register 0. Table 5-3 gives initial factory settings.  
9. Factory setting. Powers up to last stored value.  
Order of Execution  
When you send a set of instructions to the power supply, they are executed in the order in which they are received. The  
power supply completes the execution of the present command before executing another command. To send more than one  
command within the power supply command string, use a semicolon to separate the commands. This maximizes the rate at  
which the power supply accepts commands.  
Terminators  
Terminators mark the end of a command string. As shown in Figure 5-2, the semicolon, line feed < LF >, and carriage  
return line feed < CR > < LF > are the characters that indicate the end of a message to the power supply. When you are  
using the Agilent Series 200 computer with BASIC to send a command using the standard format (see Figure 5-1), the  
computer automatically sends < CR > < LF > on the data bus following the command.  
70 Remote Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Initial Conditions  
Immediately after power on from the factory, the power supply automatically undergoes a self-test and sets all parameters  
to the values contained in Table 5-3. The values in the first part of the table come from storage register 0 and were stored at  
the factory. They will remain until STO 0 is used to change settings of the power on state. When the power supply is again  
turned on, it will initialize its outputs to the values in storage register 0.  
Table 5-3. Initial Conditions  
Parameter  
Voltage  
Current  
Reprogramming Delay  
Store/Recall registers:  
0-10: Voltage  
Inital Value  
0 V High Range  
10 mA High Range  
20 mS  
0 V High Range  
10 mA High Range  
55 V (High Range)  
Disabled  
Current  
Over Voltage (OV)  
Over Current Protection (OCP)  
UNMASK register  
0 (cleared)  
SRQ  
0 (Off)  
Output Channels  
Front Panel Metering  
Power Supply Address  
Local Control  
PON bit  
On  
Output #1  
Last stored value (Factory set to 5)  
Enabled  
On  
PON SRQ  
Cal Mode  
Off (0)  
Off  
Power Supply Commands  
This section discusses the commands which you will use to program the supplys voltage and current, protection circuits,  
and enhanced features like storage and recall registers, and reprogramming delay. When programming, you should be  
aware that the current, voltage and overvoltage ranges for each output of your supply may differ. Table 5-1 shows these  
values for the power supply. If you send values out of these ranges, you will get a number range error. A summary of all  
commands appears in Appendix C.  
The output voltage of some output channels exceeds the safe operating limit of 42.2 V. To avoid any  
electrical shock, program the voltage to zero volts or turn off ac input power before changing any rear  
panel connections. Make certain all straps are properly connected, terminal block screws are securely  
tightened and terminal block covers are replaced before reapplying power.  
Voltage Programming  
To program voltage, send the output channel and the programmed value. In the example below, output l is programmed to 5  
V.  
VSET 1,5  
The values you send must always be volts. For example if you want to program 450 millivolts, convert to volts and then  
send the command:  
Remote Operation 71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
VSET 1,.45  
If the output channel is operating in constant voltage mode (CV annunciator on) then the actual voltage is the programmed  
voltage, but in CC mode of operation (CC annunciator on), the programmed voltage is the voltage limit for that output.  
When programming a value of voltage, the current setpoint will be changed if the total power is greater than the power  
boundary (see Figure 4-2).  
Example:  
ISET 4,1.5 sets current to 1.5 A  
VSET 4,50 sets voltage to 50 V  
Sending the VSET command caused the current ISET to be reduced to 1.03 A. This is an example of power limiting  
changing a supply setting. When this occurs, the "couple parameter" bit (see Table 5-5) in the status register is set.  
Table 5-4. Programmable Output Range  
Model  
Output Configuration  
Operating  
Range *  
Output Volt  
(Avg. Res.)  
Output Curr **  
(Avg. Res.)  
6625A / 6628A  
6626A / 6629A  
OUTPUT  
OUTPUTS  
Low  
High  
0-7 V  
(0.460 mV)  
0-50 V  
0-15 mA  
(1 µA)  
0-0.5 A  
(33 µA)  
1
1 & 2  
(3.2 mV)  
OUTPUT  
OUTPUTS  
Low  
High  
0-16 V  
(1.0 mV)  
0-50 V  
0.2 A  
(13 µA)  
0-2 A  
2
3-4  
(1,2-6628A)  
1,2,3,4-6629A)  
(3.2 mV)  
(131 µA)  
*
The maximum programmable voltage values for each range are 1% higher than the rated voltage and the  
maximum programmable current values for each range are 3% higher than the rated current.  
** Each channel wakes up to values stored in state register 0. The factory set values are shown in the first  
part of Table 5-3. All other registers are set to these settings. Volatile registers 4-10 always power-up to  
these conditions.  
To readback the programmed voltage setting for output 1, send the query:  
VSET? 1  
and address the supply to talk. If you want to know the value of the actual output voltage of output 1, send the query:  
VOUT? 1  
The results are placed on the GP-IB and read into the controller when the supply is addressed to talk.  
Note  
The power supply will round the VSET and ISET settings to the nearest multiple of their resolution.  
Table 5-4 lists the average resolution of these settings.  
72 Remote Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Current Programming  
To program the current, send the output channel and the programmed value in amps. In the example below, output 1 is  
programmed to 15 mA.  
ISET 1,0.015  
The value you send must always be in amps. For example if you want to program 95 milliamps, convert to amps and then  
send the command  
ISET 1,.095  
If the output channel is in constant current (CC) mode of operation, then the actual current is the programmed current but if  
the output is in the CV mode, the programmed current is the current limit of that output.  
When programming a value of current, the voltage set point can be changed if the total power is greater then the power  
boundary (see Figure 4-2).  
Example:  
VSET 4,50 Voltage set to 50 V  
ISET 4,2 ISET to 2 A  
Sending the ISET command caused the VSET to be reduced to 16.16 V. This is an example where power limiting changes  
a supply setting. When this occurs, the "coupled parameter" bit (see Table 5-5) in the status register is set.  
To readback the programmed current for output 1, send the query and addressing the supply to talk.  
ISET?  
You can also instruct the supply to measure the actual output current at output channel 1 by sending the following query  
and address the supply to talk.  
IOUT? 1  
The results are placed on the GP-IB and read into the controller.  
Range Programming  
The range can be set by using the VRSET or IRSET commands. The power supply will automatically pick the range that  
the value sent will fit into.  
Examples:  
VRSET 1,7 will select the 7 V range  
VRSET 1,3.2 will select the 7 V range  
VRSET 1,9.0 will select the 50 V range  
VRSET 1,50.5 will select the 50 V range  
IRSET 1,.015 will select the 15 mA range  
IRSET 1,0 will select the 15 mA range  
IRSET 1,.020 will select the 0.5 A range  
IRSET 1,0.1 will select the 0.5 A range  
Note  
Ranges will accept values up to 1% greater then full scale for voltage and 3 % for current. Larger  
numbers will generate a number range error(s).  
When going from the high range to the low range, the present setting must be within the boundary of the low range or the  
setting will be reduced to the maximum of the low range. If the setting is changed the "couple parameter" bit (see Table  
5-5) in the status register will be set to indicate the change.  
Remote Operation 73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Output On/Off  
The OUT command disables/enables an output channel of the power supply. It will not disturb any other programmed  
function nor will it reset the protection circuits. You can control individual outputs with the OUT command as shown  
below. For example, to disable output channel 1 send the following:  
OUT 1,0  
To enable output channel 1 send the following command  
OUT 1,1  
You can find out the present state of output 1 by sending the query:  
OUT? 1  
and addressing the supply to talk. The response from the supply is either a "0" to indicate output 1 is off or a "1" to indicate  
that the output is on. When disabled, the output behaves as if it were programmed to zero volts on the high voltage range  
and 10 mA on the high current range.  
Overvoltage (OV) Protection  
The programmable OV is a protection feature which can be set by the operator to protect the load against excessive voltage.  
When the actual voltage exceeds the programmed overvoltage setting for a given output channel, the OV is tripped. The  
OV circuit will fire the SCR crowbar which shorts across the output and the output assumes a low impedance state.  
For example, to program the OV of output channel 1 to 9.5 V send the following command:  
OVSET 1,9.5  
To find out the OV setting for output channel 1 send the following query and address the supply to talk:  
OVSET? 1  
To enable an output after it went into overvoltage, you must first remove the overvoltage condition and then send the OV  
reset command.  
To reset output 1 send:  
OVRST 1  
If you send the reset command without first removing the OV condition, the supply will fire the OV again.  
Note  
If the programmable OV fails, the supply has a fixed OV circuit which will fire the SCR crowbar if the  
voltage exceeds 120% of the maximum rated output. The fixed OV circuit will also fire the SCR crowbar  
if the supply is off (line cord disconnected) and an external source is supplying voltage which exceeds  
120% of the maximum rated output.  
74 Remote Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Overcurrent Protection (OCP)  
The OCP is a protection feature employed by the power supply to guard against excessive output currents. When the output  
enters the + CC mode and the OCP is enabled, the OCP circuit down programs the output voltage and disables the output.  
To enable the OCP, for output channel 1, send the command  
OCP 1,1  
To disable the OCP, send the command  
OCP 1,0  
You can find out the OCP setting by sending the following query and addressing the power supply to talk.  
OCP? 1  
The response from the power supply is either a "0to indicate that OCP is off or a 1" to indicate that it is on. To reset the  
output channel after an overcurrent trip, you can either disable the OCP and send the reset command, or you can reduce the  
output current below the programmed current and then send the reset command. To reset output 1, send the command:  
OCRST 1  
Note  
The supply can report a fault condition when an output is in overvoltage or overcurrent. Although the  
OVRST and OCRST commands re-enable the output, they do not clear the fault register. As a  
housekeeping measure, it is advisable to always clear the fault register by querying its value after an OV  
or OC reset.  
Multiple Output Storage & Recall  
The power supply has 11 internal registers (0-10) each of which can store the voltage and current settings of all the outputs.  
By storing voltage, voltage range, current, current range and OV settings for all outputs and recalling them later, you can  
have significant savings in programming time. (See Supplemental Characteristics in Table 1-2). Upon shipment from the  
factory, registers (1-10) contain values of 0 volts and 0 current (see Table 5-4).  
Registers 0-3 are non-volatile. They are limited to one STORE each. If more than one STORE is attempted, error 30  
(STORE LIMIT) will result. The line must be turned off/on to re-enable these store locations. Register 0 is a special case  
that also contains Delay, Mask. When the power supply is turned on, it will access register 0 for power-up settings.  
Registers 4-10 (volatile) will be reinitialized to 0 volts and 10 mA (max OV & V, I ranges = HIGH) when power is cycled  
on/off.  
To store voltage and current settings, issue the store (STO) command and specify a register (0-10). For example, to store  
the present setting of current and voltage of all outputs in register 2, send the following command:  
STO 2  
To set the power supply to the voltage, voltage range, current, current range, and overvoltage settings stored in a register,  
issue the recall (RCL) command followed by the desired register. For example to recall settings stored in register 2, send  
the following command:  
RCL 2  
When a register is recalled, the outputs will be set sequentially (output 1, output 2, etc.). If you attempt to recall a register  
outside the 0 to 10 range, you will get a number range error (GP-IB code 5 in Table 5-8).  
Remote Operation 75  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The Clear Command  
This command will return the power supply to its power-on state and all parameters are returned to their initial power-on  
values except for the following:  
1. The store/recall registers are not cleared.  
2. The power supply remains addressed to listen.  
3. The PON bit in the serial poll register is cleared  
To clear the power supply, send the following command:  
CLR  
Status Reporting  
The power supply has the ability to report its internal status to the user whenever it is asked to do so. Depending on the type  
of status the user requested, the supply will interrogate the status, accumulated status, mask, or fault registers present in  
each output. The status register can report status independently (see page 76) or it can work together with the mask and  
fault registers to report a fault (see page 77). The accumulated status register records every status condition the output  
experienced since the time it was last read (see page 77). Figure 5-3 shows a conceptual model of the operation of these  
registers.  
Figure 5-3. Functional Relationship of Status Registers  
The supply has one serial poll register which services all outputs and provides the user with other power supply status-  
related information as discussed in page 78.  
Status Register. Each output channel of the power supply maintains its present status in an eight bit register. This status  
register reports the status of the output channel whenever it is queried. A "1" in any of the bit positions indicates that the  
condition is true. As long as the condition continues to be true, the bit will remain set. Assignments for the bits are shown in  
Table 5-5.  
Decoding of the reading is based on the weighted number placed on each bit of the eight bit status registers. For example,  
bit position 5 in the register has a bit weight of 32 (see Table 5-5). Each bit is assigned to a particular condition and the  
corresponding bit weight is used to identify that condition. When set, bit 5 indicates that the associated output is in the  
unregulated state. If this is the only bit that is set, the number 32 will be returned when the outputs status register is  
queried.  
76 Remote Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 5-5. Bit Assignment for the Status, Astatus, Fault, and Mask Registers  
Bit Position  
Bit Weight  
Meaning  
Where  
7
128  
CP  
6
64  
OC  
5
32  
UNR  
4
16  
OT  
3
8
OV  
2
4
-CC  
1
2
+CC  
0
1
CV  
CV =  
+ CC =  
- CC =  
OV =  
Constant Voltage Mode  
Positive Constant Current Mode  
Negative Current Limit Mode  
Overvoltage Protection circuit tripped  
OT =  
UNR =  
OC =  
Over Temperature Protection circuit tripped  
Unregulated Mode  
Over Current Protection tripped  
Coupled parameter (See Note)  
CP =  
Note: When the range is switched as discussed on page 73, or the power limit is exceeded (see pages 71 &  
73), the CP bit is set. It is cleared when you send a voltage, current or range value that causes no other  
changes.  
To query an output channel for its status, you must specify the output channel. For example, to find out the status at output  
2 send the following query and address the supply to talk:  
STS? 2  
Accumulated Status Register. Each output channel of the power supply also maintains a cumulative status in its  
accumulated status (astatus) register. This register records every status condition the power supply output entered since it  
was last queried. When queried, it returns a decimal number which is decoded as shown below. The astatus register is reset  
to the present value of the status register after it is queried. The bits are assigned as in Table 5-5. Here is an example to help  
you decode the decimal number (from 0 to 255) returned when the astatus register is queried. If the output channel was in  
overvoltage since the last reading of the astatus register and that channel is presently operating in constant voltage mode,  
the reading you will get when you query the register will be 9. To decode this we use Table 5-5.  
9
= 8  
+
1
OV  
+
CV  
For example, to query the astatus register of output 2, send the following query and address the supply to talk  
ASTS? 2  
The Mask and Fault Register. The fault register works in conjunction with the mask register. These are two eight bit  
registers which report any fault condition on a particular output channel. The mask register is used to set up the conditions  
that generate a fault which is latched into the fault register. The user can then read the fault register to determine the fault.  
When a bit in the fault register is set, the power supply can generate a service request for that output providing the service  
request command on fault (SRQ 1 or SRQ 3) was previously sent. See page 79 for a discussion on service request.  
To understand how these two registers work, we must include the status register in this discussion. Recall that the status  
register takes its input from the power supply and the user cannot change its contents. The mask register takes its inputs  
from the user, and the power supply cannot change its contents. The fault register takes its inputs from both the mask and  
the status registers. You can find out the setting of the mask register of output 2 by sending the following query and  
addressing the supply to talk:  
UNMASK? 2  
The response will be a numeric code between 0 and 255 which can be decoded by consulting Table 5-5. You can set the  
conditions to generate a fault by setting (unmasking) one or more bits in the mask register. The conditions will remain  
unmasked until you change them. To unmask conditions in output 2 for example, send the following command:  
Remote Operation 77  
Download from Www.Somanuals.com. All Manuals Search And Download.  
UNMASK 2,XXX  
where XXX specifies the numeric code (0 to 255) for the unmasked conditions (see Table 5-5). If during operation, the  
output experiences any of the previously unmasked conditions, it will set the corresponding bit(s) in its fault register.  
Remember that the bits in the fault register can be set when there is a change in either the status register or the mask  
register. Each output has its status, mask, and fault registers arranged as shown in Figure 5-3 and Table 5-5. The mask  
register, which is set by the user, is used to specify which bits in the status register are enabled (unmasked) to set bits in the  
fault register. A bit is set in the fault register when the corresponding bit in the status register changes from 0to 1 a n d  
the corresponding bit in the mask register is a "1". Also, if a bit in the status register is already set and then the  
corresponding bit in the mask register is set (unmasked), the corresponding bit in the fault register will be set.  
In addition, if both status and mask register bits remain set after the fault register was read (and cleared), the fault register  
will remain cleared as long as there are no changes in either the status or mask registers with the following exception.  
Executing a VSET, ISET, RCL, OVRST, OCRST, or OUT on/off command, will cause the CV, + CC, - CC, or UNR bit  
(as applicable) in the fault register to be set. Note that the fault register is cleared immediately after it is read.  
As shown in Figure 5-3, if one or more bits in the fault register of a given output channel are set, then the FAU bit for that  
output in the serial poll register will also be set and a service request may be generated (see page 79). To read the fault  
register of output 2 and find out which bits are set, send the following query and address the supply to talk:  
FAULT? 2  
The power supply responds with a number which can be decoded from Table 5-5. For example, the number 9 (8 + 1)  
indicates that the OV and the CV bits in the fault register are set.  
Note  
If the condition(s) generating the fault(s) is (are) removed but the fault register is not read, the bit(s) in the  
fault register will remain set.  
The Serial Poll Register. The serial poll register is an 8 bit register which the supply uses to keep track of its internal  
operating status and to determine the operating status of each of its outputs. Table 5-6 defines each bit.  
Table 5-6. Bit assignment of the Serial Poll Register  
Bit Position  
Bit Weight  
Meaning  
7
6
64  
RQS  
5
32  
ERR  
4
16  
RDY  
3
8
2
4
1
2
0
1
128  
PON  
FAU 4  
FAU 3  
FAU 2  
FAU 1  
The first four bits (0 to 3) in the register tell whether or not a particular output has a fault. If there is a fault in one of the  
outputs, then the corresponding FAU bit will be set. Thus if output 1 has a fault, then FAU 1 will be set. In models with  
only three outputs, FAU 4 will always be zero and in two output models, FAU 3 and FAU 4 will always be zero.  
The RDY bit is set when processing is complete and is cleared when the supply is processing commands.  
The ERR bit is set when a programming or hardware error occurs and is cleared when the error query (ERR?) is received.  
The error annunciator on the front panel informs the user when this bit is set or cleared.  
The RQS bit is set when the power supply generates a service request and cleared after a serial poll is done (see page 79).  
The PON bit is set at power on and cleared when a CLR command is sent.  
78 Remote Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Request Generation  
When operating your supply, you may want it to request service every time a fault or a programming error condition  
occurs. To do this you send a service request (SRQ) command. When the condition is true, the power supply responds by  
setting the RQS bit in the serial poll register, setting the SRQ annunciator on the front panel, and issuing an SRQ over the  
GP-IB.  
The power supply can generate a service request for any of the following reasons: (refer to Table 5-7)  
An Output Fault. If there is a fault on one or more of the output channels and you previously sent the SRQ 1 or  
SRQ 3 command (see Service Request Enable/Disable information below), then an SRQ will be generated.  
An Error. If there is an error (see Tables 5-8) and you previously sent the SRQ 2 or SRQ 3 command, (see  
Service Request Enable/Disable information below), then the supply will generate a service request.  
Power-on. At power-on, the PON bit of the serial poll register is set but the supply will only generate an SRQ if  
you previously sent a PON 1 command.  
Input Line Voltage Dropout. Same as power-on condition.  
To find out the nature of the service request, you must do a serial poll. This will isolate the output that generated the request  
by checking which of the FAU bits are set in the case of a fault, or checking to see if the error bit is set in the case of an  
error. If the SRQ on faults was set, then send the fault query.  
FAULT? 2 (using output 2 as an example)  
and address the supply to talk if you want to find out which of the conditions you unmasked in Figure 5-3 are true. For  
example if the supply was in overvoltage and that condition was unmasked then the response from the fault query will be  
"8(see Table 5-5).  
Note  
When you query the fault, the fault register is cleared. Performing a serial poll will reset the RQS bit but  
will not clear the fault register.  
If the SRQ on error was set then you can send the error query ERR? and address the supply to talk. The response will  
identify the error by its code (see Table 5-8).  
Service Request Enable/Disable. You can query the status of the service request enable/disable function by sending  
the query:  
SRQ?  
and addressing the power supply to talk. The response from the supply is one of the following:  
0, 1, 2, or 3  
0--indicates that the service request capability (except for power-on; see page 80) is disabled.  
1--indicates that it is enabled for output fault conditions.  
2--indicates that it is enabled for error conditions.  
3--indicates that it is enabled for both fault and error conditions.  
Remote Operation 79  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The ability to generate service requests can be enabled or disabled using the SRQ command as described below.  
To disable the service request capability, except for power-on, send:  
SRQ 0  
To enable the service request capability for all output faults, send:  
SRQ 1  
To enable the service request capability for errors, send:  
SRQ 2  
To enable the service request capability for both faults and errors, send:  
SRQ 3  
The Power-On Service Request. You can also cause the power supply to request service every time it is switched on  
or every time there is a temporary loss in power. To do this send the following command:  
PON 1  
If you want to disable this facility, send the command.  
PON 0  
If you want to find out if the power-on SRQ is enabled or disabled, send the following query:  
PON?  
and address the supply to talk. The supply will respond with a 1 or 0 as discussed above.  
Note  
The power-on (PON) SRQ mode is stored in the non-volatile memory of the supply so that although the  
supply may be switched off, it will remember the status of the last PON command at power-on and  
respond accordingly.  
Table 5-7 summarizes all the conditions under which a service request will be generated.  
Table 5-7. Conditions for Generating a Service Request  
Condition  
Commands Sent  
State of  
PON  
SRQ  
0
-
2 or 3  
1 or 3  
RQS Bit  
0
1
-
0
1
1
1
Any  
Power-on  
Error  
-
Fault  
Reprogramming Delay  
The power supply may switch modes or become unregulated momentarily after a new output value is programmed.  
Because of their short duration, these cases may not ordinarily be considered a fault but the supply will recognize this  
deviation and generate a fault signal. To prevent this, the reprogramming delay feature is implemented.  
80 Remote Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Reprogramming delay will delay the onset of certain fault conditions and prevent the power supply from registering a fault  
when these conditions are true. When the delay is in effect, the CV, + CC, - CC and UNR bits of the status register are  
masked and cannot communicate with the mask and fault registers and the OCP function. This will prevent the supply from  
registering a fault should any of these bits become set during the delay period. Reprogramming delay is initiated when any  
of the following functions are executed:  
VSET, ISET, RCL, OVRST, OCRST, OUT on/off  
At power-on reprogramming delay is set to 20 mS. You can specify new values between 0 and 32 S in steps of 4 mS. If you  
specify a value which is not a multiple of 4 mS, the supply will round off the set value to the nearest 4 mS multiple.  
To program a new value of 80 mS in output 2 for example, send the following:  
DLY 2,.08  
If you send a value outside the 0 to 32 S range you will get a programming error. You may use the programming response  
times in the specifications table to give you an idea of a typical delay setting. However, the appropriate delay setting will  
also depend on load capacitance, load resistance, and current limit setting. See page 53 for output capacitor considerations.  
To query the reprogramming delay setting of a particular output channel, send the following query:  
DLY? 2 (using output 2 as an example)  
and address the supply to talk. The response will be a numeric value between 0 and 32.  
Display On/Off  
When the display is on, the commands sent across the GP-IB may experience a slower processing time because the  
processor must also spend time to monitor the outputs and update the display. You can shorten your command processing  
time by turning off the display. To turn off the display, send the command:  
DSP 0  
To re-enable the display send the command:  
DSP 1  
You can also find out the status of the display by sending the following query and addressing the supply to talk:  
DSP?  
The response will be either a "1or a "0’.  
Message Display Capability. The display command can also be used to display messages on the front panel. Messages  
may consist of a maximum of twelve alphanumeric characters. Only upper case alpha characters, numbers, and spaces will  
be displayed.  
For example, to display the message "OUTPUT 2 OK", send the following command:  
DSP "OUTPUT 2 OK’  
Note  
The BASIC programming statement for a series 200/300 computer would be as follows:  
OUTPUT 705; "DSP""OUTPUT 2 OK"""  
Remote Operation 81  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Other Queries  
In the examples discussed above, you saw how to use queries for each function discussed. The following paragraphs  
describe other queries which were not previously covered.  
ERROR Query. The power supply can detect both programming and hardware errors. You can use either the front panel  
(see page 89) or the GP-IB to find out the type of error. Upon detecting an error, the error annunciator on the front panel  
and the ERR bit in the serial poll register will be set. When in local mode the supply will display the error name in response  
to pressing the ERR key on the front panel. Over the GP-IB, only the error code will be returned. After a query, the error bit  
is cleared. A description of these codes is given in Tables 5-8. To find out what the error is, send the following query and  
address the supply to talk:  
ERR?  
The supply will respond with an error code number (see Table 5-8).  
ID Query. If you want to know the model number of the power supply you are working with, you can send the ID? query  
over the GP-IB. To do this send the following over the GP-IB and address the supply to talk.  
ID?  
The supply will respond with its model number.  
Test Query. You can get the power supply to perform a limited self-test at any time during its operation by sending the  
TEST? query over the GP-IB. This test does not affect the analog control circuits of the supply and it can be performed  
while the outputs are connected to external circuits. For tests of the analog control circuits refer to Chapter 3. Responses to  
the test query are described in Table 5-9. This test cannot be done from the front panel. To instruct the power supply to  
carry out a self-test, send the following query and address the supply to talk:  
TEST?  
Calibration Mode Query. To be able to calibrate your power supply, the calibration mode (CMODE) must be turned on  
(See Appendix A for a detailed description of the calibration procedure). To find out if the CMODE is on or off, send the  
following query over the GP-IB and address the supply to talk:  
CMODE?  
The supply will respond with a 1" which indicates that CMODE. is on or a "0" which indicates that CMODE is off.  
DCPON. The DCPON command sets the state of all outputs at power-on. You can specify if the outputs wake up enabled  
or disabled when the unit is turned on. To enable all outputs at turn-on send:  
DCPON 1  
To disable all outputs at turn-on send:  
DCPON 0  
Note that these commands set the output voltage to zero and the output current to a slightly positive value when the  
OUTPUT OFF command is issued. Therefore, the constant voltage feedback loop is active and the outputs are in constant  
voltage mode when programmed “OFF”.  
Starting with firmware revision A.00.04 and later, two additional commands let you set the output current to a slightly  
negative value when the OUPUT OFF command is issued, causing the outputs to be in constant current mode when  
programmed “OFF”. The benefit of being in constant current mode when the output is off is that if the load impedance and  
the voltage and current settings are such that the unit is forced into constant current mode at turn on, then the current  
feedback loop will be active during the transition to OUTPUT ON, and there will be no output current overshoot due to  
mode crossover.  
82 Remote Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
To enable all outputs in constant current mode at turn on send:  
To disable all outputs in constant current mode at turn-on send:  
DCPON 2  
DCPON 3  
Table 5-8. Error Messages  
Explanation  
Front Panel  
Response  
NO ERROR  
GP-IB  
Code  
0
This is the response to the ERR? query when there are no errors.  
You sent the supply a character it did not recognize.  
INVALID CHAR  
INVALID NUM  
INVALID STR  
1
2
3 or 28  
4
Format of your number is incorrect. Check number syntax.  
Occurs when you send a command the supply does not understand.  
SYNTAX ERROR  
Either too many parameters are sent without delimiters or the number  
representation is incorrect. Follow the Syntax Diagram in Figure 5-2. Check spaces  
and delimiters.  
NUMBER RANGE  
NO QUERY  
5
6
An out of range number was sent. Send a new number within the legal range.  
Computer addressed the supply to talk, but it did not first request data. Send query  
first then address supply to talk.  
DISP LENGTH  
7
Quoted string exceeds the display length of 12 characters. Shorten string to a  
maximum of 12 characters.  
BUFFER FULL  
8
9
This error may occur if too many numbers are sent. Error #4 or #5 may occur first.  
EEPROM ERROR  
EEPROM is not responding correctly to programming commands. An instrument  
failure has occurred and service is required.  
HARDWARE ERR  
10  
An output error has occurred in an unknown output. Service is required.  
Remote Operation 83  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 5-8. Error Messages (continued)  
Explanation  
Front Panel  
Response  
HDW ERR CH 1  
GP-IB  
Code  
11  
Errors 11 through 14 refer to a specific output where there is an output error.  
Service is required.  
HDW ERR CH 2  
HDW ERR CH 3  
HDW ERR CH 4  
NO MODEL NUM  
12  
13  
14  
15  
Same as in Error #11.  
Same as in Error #11.  
Same as in Error #11.  
The interface cannot find its model number. There may be a hardware failure or  
the instrument may require reprogramming. Service is required.  
CAL ERROR  
16  
17  
You tried to use either a calibration command with CMODE off or the calibration  
failed while in CMODE. Enable CMODE and check numbers sent during  
calibration. Also, there could be a hardware error.  
UNCALIBRATED  
There is an incorrect checksum in the EEPROM possibly as a result of incorrect  
calibration procedure. Recalibrate, and if the problem persists, your supply has a  
hardware failure.  
CAL LOCKED  
SKIP SELF TST  
18  
22  
Calibration was attempted with the Calibration Jumper on the GP-IB board in the  
lockout position. Reposition jumper if desired. See Service Manual.  
The self test jumper on the GP-IB board is in the Skip Self Test position. No self-  
test was done. This is for diagnostics only. See Service Manual.  
Table 5-9. TEST Responses  
Explanation  
Code  
0
This is the response to the TEST? query when there are no errors.  
20  
21  
27  
29  
The timer has failed self-test. Refer to the troubleshooting section in the Service Manual.  
The RAM has failed self-test. Refer to the troubleshooting section in the Service Manual.  
The ROM has failed the checksum test. Refer to the troubleshooting section in the Service Manual.  
There is an incorrect checksum in the EEPROM for non-volatile state storage. The non-volatile states 0-3  
have not been loaded. Store new values, if problem persists, your supply has a hardware failure.  
30  
The store limit has been reached for CMODE, store states 0-3, DCPON, or SRQ. The power must be  
turned off/on to reenable.  
84 Remote Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6
Local Operation  
Introduction  
Chapter 3 introduced you to the supplys front panel controls and indicators to help you turn on the supply and perform the  
checkout procedures that were given in that chapter. The following paragraphs describe how to use all of the front panel  
controls and indicators. Most of the remote operations described in Chapter 5 can also be performed locally from the  
supplys front panel.  
Local Mode  
In order to use the front panel keys to control the supply, the local mode must be in effect. The local mode is in effect  
immediately after power is applied. Table 5-3 lists the initial settings for all of the power supplys functions when power is  
initially applied. When the local mode is in effect (RMT annunciator off), the Output Function, System Function, and  
Numeric Entry keys on the front panel (see Figure 6-1) can be used to operate the power supply.  
In the remote mode (front panel RMT annunciator on), the front panel keys will have no effect on any of the supplys  
outputs and only the computer can control the supply. You can, however, still use the front panel display to view the output  
voltage and current readings or the present settings for the selected output channel while the supply is in the remote mode.  
You can return the supply to the local mode from the remote mode by pressing the LCL key provided that the local lockout  
command has not been received from the GP-IB controller. Pressing the LCL key will also turn the supplys display back  
on if it was turned off with a DSP command during remote operation (see page 81). A change between the local and remote  
modes will not result in a change in the power supplys outputs.  
Local Control Of Output Functions  
The Output Function keys (see Figure 6-1) allow you to control the selected output. Figure 6-1 shows the annunciator arrow  
over OUTPUT 2 indicating that output channel 2 is selected. Pressing the OUTPUT SELECT key selects the output  
channels in forward (>) or reverse (<) sequence. Note that Figure 6-1 illustrates the front panel for the Agilent 6626A  
supply which has four output annunciators. The front panel for Agilent Model 6625 is identical to Figure 6-1 except it has  
two output annunciators.  
General  
The power supply will accept programming values directly in volts and amps. The programmable voltage, current, and  
overvoltage ranges for the outputs of each model are given in Table 5-4. The power supply will round off the values  
received to the nearest multiple of the resolution for that particular output. If you send a value out of the valid range, it will  
not be executed and the ERR annunciator will come on. You can get a readout of the error on the display by pressing the  
ERR key. For an out-of-range error, the error message NUMBER RANGE" will be displayed.  
When you press the VSET, ISET, or OVSET key, the output selected and the present setting for that function will be  
displayed. You can change the setting using the numeric entry keys. Pressing the number keys will cause the present  
numeric setting to become blank and be replaced with the new numbers on the display. You can use the Ï key to erase  
previous keystrokes if you make a mistake.  
Local Operation 85  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LOCAL MODE  
KEY  
OUTPUT  
ANNUNCIATOR  
CHANNEL 1  
DISPLAY  
OUTPUT FUNCTION KEYS  
6626A SYSTEM DC POWER SUPPLY  
SYSTEM  
OUTPUT  
ENTRY  
VOLTS  
AMPS  
LCL  
OUTPUT  
SELECT  
ADDR  
ERR  
STO  
RCL  
RANGE  
V/I  
7
4
1
0
8
5
2
.
9
6
3
VOLT  
VOLT  
45.153 14.235m  
OVSET  
OCP  
VSET  
ISET  
1
2
3
4
CV CC UNR OCP ERR RMT ADDR SRQ  
ENBLD  
-- OUTPUT --  
ENTER  
CURR  
CURR  
OUTPUT  
ON/OFF  
RESET  
LINE  
ON  
OFF  
SYSTEM FUNCTION KEYS  
NUMERIC ENTRY KEYS  
Figure 6-1. Front Panel (Model 6626A shown)  
Pressing the ENTER key will enter the values displayed for the function indicated, initiate that function, and return the  
display to the metering mode in which the measured output voltage and current for the selected output are displayed.  
Pressing the ENTER key without entering numbers will result in retention of the previous values and return to the metering  
mode. You can also return to the metering mode at anytime by pressing the RESET key.  
The RANGE V/I key is used to display the voltage and current range settings. The key toggles; pressing it will display the  
range, pressing it again will put the display back to background metering. The range display shows the largest values of  
voltage and current that can be set for the ranges presently set.  
The voltage step and current step (arrow keys) are used to change voltage & current settings and voltage & current ranges.  
There are 3 modes:  
1. To use the step keys to immediately change the output settings, the background metering mode must be in affect.  
2. To use the step keys to change the setting, the VSET or ISET key must be pressed first. The step keys can then be used to  
adjust the setting. Press the ENTER key to change the actual setting.  
3. To use the step keys to change the voltage or current ranges first press the RANGE V/I key. The step keys change the  
range as long as the RANGE display is in effect.  
Note  
The step keys change the voltage or current by 1 LSB each time they are pressed. If the key is held down  
more than 1 second, the automatic step mode goes into effect. The step rate will start to increase from the  
initial value after about 3 seconds.  
Setting Voltage and Voltage Range  
The selected outputs voltage is programmed locally using the VSET key. For example, program the voltage to 5.25 volts  
by pressing:  
VSET  
5
.
2
5
ENTER  
The front panel display then indicates the actual output voltage and current for the selected output.  
86 Local Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The voltage step keys can also be used to change the voltage setting.  
VSET ÑVOLT or ÒVOLT  
then  
ENTER  
The voltage can be set in the immediate execute mode. This mode is in effect when background  
metering is in effect. Pressing ÑVOLT or ÒVOLT will change the setting  
and the effect on the voltage and current can be observed on the display.  
The voltage range can be changed by using the RANGE V/I key and the voltage step keys.  
RANGE V/I ÑVOLT or ÒVOLT  
The range will change as soon as one of the voltage step keys is pressed and the display will be updated. If the voltage  
setting is greater then the highest value of the low range and the range is changed from high to low, the VSET will be  
automatically reduced to the maximum voltage of the low range.  
Setting Current and Current Range  
The selected outputs current is programmed locally using the ISET key. For example, program the current to 1.5 amps by  
pressing:  
ISET  
1
.
5
ENTER  
The power supply will accept any programmed current between zero and the minimum programmable current and  
automatically set the output to the minimum programmable current without causing a programming error. See Table 5-4.  
The current step keys can also be used to change the current setting.  
ISET  
ÑCURR  
or  
ÒCURR  
then  
ENTER  
The current can be set in the immediate execute mode. This mode is in effect when background  
metering is in effect. Pressing ÑCURR or ÒCURR will change the setting  
and the effect on the voltage and current can be observed on the display.  
The current range can be changed by using the RANGE V/I key and the current step keys.  
RANGE V/I ÑCURR or ÒCURR  
The range will change as soon as one of the current step keys is pressed and the display will be updated. If the current  
setting is greater then the highest value of the low range and the range is changed from high to low, the ISET will be  
automatically reduced to the maximum current of the low range.  
Enabling/Disabling an Output  
The selected output channel can be turned on and off from the front panel. The OUTPUT ON/OFF key toggles the selected  
output on and off. When an output is turned off, the message "DISABLED" will be displayed.  
The OUTPUT ON/OFF key will not affect any other programmed functions nor will it reset an overvoltage or overcurrent  
condition. An output disabled by the OUTPUT ON/OFF key will behave as if it were programmed to zero volts and 10 mA.  
Local Operation 87  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Setting Overvoltage Protection  
Programmable overvoltage protection (OVP) guards your load against overvoltage by crowbarring and downprogramming  
the power supply output if the programmed overvoltage setting is exceeded.  
A fixed OV circuit with a trip level about 20 percent above the maximum programmable voltage acts as a backup to the  
programmable OVP. When overvoltage protection is activated, the output is shorted and the message OVERVOLTAGE’  
will appear on the front panel display.  
The selected outputs overvoltage setting is programmed locally using the OVSET key. For example, program the  
overvoltage to 10.5 volts by pressing:  
OVSET  
1
0
.
5
ENTER  
Resetting Overvoltage Protection  
The condition that caused the OVP must first be cleared and then the output can be returned to its previous state by pressing  
the RESET key.  
Enabling/Disabling Overcurrent Protection  
The overcurrent protection feature guards against excessive output currents. When the output goes into the + CC mode and  
OCP is enabled, the OCP circuit is activated which downprograms the output voltage and disables the output. For this  
condition, the message OVERCURRENTappears on the front panel display.  
The selected outputs overcurrent protection feature can be turned on and off from the front panel. The OCP key toggles the  
selected outputs overcurrent protection circuit on and off. When it is on (enabled), the OCP ENBLD annunciator will be  
on.  
Resetting Overcurrent Protection  
The condition that activated the OCP circuit must first be cleared and then the output can be returned to its previous state by  
pressing the RESET key.  
Local Control Of System Functions  
The System Function keys consist of the ADDR, ERR, STO, and RCL keys as shown in Figure 6-1. These keys are  
independent of the output selected and are used in setting the supplys GP-IB address, displaying error messages, and  
storing/recalling voltage and current settings for all of the supplys output channels.  
Setting the Supply’s GP-IB Address  
As described in page 40, before you can operate the supply remotely, you must know its GP-IB address. You can find this  
out locally from the front panel by pressing:  
ADDR  
The supplys present address will appear in the display. Address 5 is the factory set address.  
If you want to leave the address set at 5, you can return to the metering mode by pressing the METER key or you can press  
another function key.  
If you want to change the address, you can enter a new value.  
Any integer from 0 through 30 can be selected.  
88 Local Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
For example, you can change the address of your supply to 10 by pressing:  
ADDR  
1
0
ENTER  
Displaying Error Messages  
The power supply can detect both programming and hardware errors. Upon detecting an error, the ERR annunciator on the  
front panel comes on and the ERR bit in the serial poll register will be set (see page 78).  
When an error is detected, you can display the error message by pressing the ERR key. The power supply will return the  
error message to the display and clear the error bit. For example, if you program a number that is not within the valid range,  
the ERR annunciator will come on. You can display the error message by pressing the ERR key. In this case, the error  
message "NUMBER RANGE" will be displayed. Errors generated either locally from the front panel or remotely from the  
GP-IB computer can be displayed by pressing the ERR key only when the supply is in the local mode. Pressing the ERR  
key also clears the error bit so if you press ERR again, the message "NO ERROR" will appear. All error codes and  
associated display messages are listed in Table 5-8.  
Storing and Recalling Voltage and Current Settings for All Outputs  
As described on page 75, the power supply has 11 internal registers for storing voltage and current settings of all outputs.  
At power on each location contains zero volts and the minimum current limit (see initial conditions on page 71).  
The STO and RCL keys allow you to store and recall voltage and current settings for all your output channels from any of  
the 11 internal registers (numbered 0 through 10). For example, you can store the present settings of voltage and current for  
all the output channels in internal register 2 by pressing:  
STO  
2
ENTER  
You can change the settings of any of your supplys outputs any number of times as required and then program them to the  
settings stored in internal register 2 by pressing;  
RCL  
2
ENTER  
The internal registers 0-3 will retain the settings when power is turned off. When power is turned off and then on again,  
registers 4-10 will be reset to zero voltage and the current to 10 mA. Register 0 is the power on state. The power will be  
programmed from values in register 0 at power on.  
The advantages in using the internal registers are that command processing time is saved and repetitive programming of  
different settings is simplified. The STO key can be used in conjunction with the OUTPUT ON/OFF key to store settings  
while the outputs are disabled (OFF). These stored settings can be used later to program the outputs to the stored settings  
using the RCL and OUTPUT ON/OFF keys.  
Local Operation 89  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
A
Alignment Procedures  
Introduction  
This appendix contains the software alignment procedures for the power supply. These supplies should be aligned twice a  
year, or whenever certain repairs are made (see Service Manual).  
The equipment that you need and the test setups to perform the alignment are also shown. A listing of the alignment  
procedure appears at the end of Appendix A. Refer to Appendix C (Command Summary) for detailed explanations of the  
commands used within the alignment program.  
Because there are no internal or external hardware adjustments, your power supply can be aligned without removing the  
covers or removing it from the cabinet if it is rack mounted. Alignment is performed by measuring actual output values and  
sending them to the supply over the GP-IB.  
Security against accidental alignment is available. A jumper inside the unit may be moved from NORMAL to CAL  
LOCKOUT to disable all alignment commands. Access to this jumper requires opening the unit (see Service Manual).  
If a number returned to the power supply is out of range while running the alignment program, the front panel ERR  
annunciator will indicate an error. For example, an error will be created if the DMM is connected to measure a current  
when a voltage measurement is anticipated by the program.  
Note  
The memory used to store correction constants will accept and store data about 10,000 times, which is  
more than sufficient for normal calibrations over the life of the instrument. However, do not put the unit  
in a calibration loop that repeatedly turns the calibration mode on and off.  
Test Equipment and Setup Required  
The following test equipment is required for alignment:  
1. A computer connected to the GP-IB connector on the back of the power supply.  
2. An Agilent 3458A Digital Multimeter (DMM).  
3. A precision 0.1 ohm ±0.003%, 5 A shunt resistor (4 terminal). Recommended--Leeds & Northrop Part Number 4221-B  
(in oil bath) or Guildline - Part Number 9330/.l/Agilent (in air).  
In line 130, change the constant SHUNT_r to the actual measured value of the shunt being used.  
Figures A-1 through A-5 show setups required for aligning the power supply. You will be instructed by the program to  
refer to a particular setup. Observe polarity when connecting the DMM.  
Calibration Program  
The following alignment program can be used as is, provided you have an HP Series 200/300 computer with the BASIC  
programming language and an Agilent 3458A DMM. Refer to Appendix C (Command Summary) for further information  
on the commands used within the program.  
Alignment Procedures 91  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The program contains embedded comments (identified by a leading asterisk {! } ) which explain various sections and  
procedures. To reduce keystroking, the program may be shortened to a minimum number of lines by eliminating these  
comments.  
The alignment program is available on a 5¼ floppy (Agilent P/N 06626-10001) or 3½ inch microfloppy (Agilent P/N  
06626-10002). Both disks also contain a verification program, which tests the power supply's output programming and  
readback accuracy to ensure that it is aligned correctly.  
The program will automatically recognize what model power supply is being tested, the number of outputs the power  
supply has, and which type output it is testing. Using this information, the program will calibrate each of the power supply.  
After the program has finished calibrating the power supply, it will store the calibration constants and turn off the  
calibration mode.  
Type in the program and press RUN.  
The program first does some "house keeping".  
It will clear the screen, and assign a number of Common Variables (lines 100 - 130).  
The program next assigns the address of the power supply to 705 (line 160) and the address of the DMM to 722 (line 170).  
It will then assign the CRT as the output device (line 180) and the value of the current shunt. Enter the exact measured  
value at line 190.  
The power supply is sent a CLR command (line 210), which will set it to the turn on state. It will then identify what model  
power supply is being tested (lines 220 and 230).  
Next it will preset the DMM (line 250) and set the DMM sample (line 260).  
It will now place the power supply in calibration mode (line 300).  
The program is now ready to perform the calibrations on the power supply.  
The program will first perform a subprogram (called at line 360 and performed in lines 3420 - 3630) which will set up the  
output parameters of the output under test.  
The program will then calibrate its voltage programming and readback circuitry (the subprogram is called at lines 370 and  
380, and performed in lines 1430 - 1920). During this subprogram, the controller will pause, and request that the user set up  
the power supply and DMM referring to a figure in this Appendix. Press CONTINUE when ready.  
Next the overvoltage circuitry will be calibrated (the subprogram is called in line 390, and performed in lines 1960 - 2070).  
The current alignment tests on the 2 Amp, 50 W outputs are made using a shunt, be sure that the DMM is connected to the  
HI and LO voltage input terminals when the shunt is used.  
The current circuitry will be calibrated next (the subprograrn is called in lines 400 and 410, and performed in lines 2110 -  
2700). During this subprogram, the controller will pause and request that the user set up the power supply and DMM  
referring to a figure in this Appendix. Press CONTINUE when ready.  
Finally the program will calibrate its current sink readback circuitry (performed in lines 2740 - 3380).  
To perform the current sink readback alignment on the Agilent 6626A, outputs of equivalent current capability are used as  
pairs. For example outputs 1 and 2 are used to test each other, the outputs 3 and 4 are used to test each other.  
In order to perform the current sink readback alignment on the Agilent 6625A, the program automatically limits the current  
of output 2 (the 50 W output) to 0.5 Amps (line 2780).  
92 Alignment Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
After testing of a channel is completed, the program will check if any errors have occurred (the subprogram is called in line  
850 and performed in lines 3670 - 3780). If an error has occurred, a message will be sent to the output device (CRT).  
After all outputs are tested without errors, the CAL MODE will be turned off, and the calibration constants stored.  
Figure A-1  
Figure A-2  
Alignment Procedures 93  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Figure A-3  
Figure A-4  
Figure A-5  
94 Alignment Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alignment Program  
10 ! This program called "ALIGN_6626" will align  
20 ! the Agilent6625A, 26, 28 and 29A Power Supplies It  
30 ! requires an Agilent3458A DMM and a four terminal  
40 ! 0.1 ohm current shunt accurate to +/-50 ppm  
50 !  
60 ! May 06,l989 Rev A.03.01  
70 !  
80 Clear_screen  
! Call program to clear display  
90 !  
100 COM /Specs/ Vrng_lo,Vrng_hi,Irng_lo,Irng_hi  
110 COM /Ps/ Chan,Other_chan,@Ps,Model$[7]  
120 COM /Outputs/ No_of_outputs  
130 COM /Vm/ @Vm  
140 COM /Shunt/ Shunt_r  
150 !  
160 ASSIGN @Ps TO 705  
170 ASSIGN @Vm TO 722  
180 PRINTER IS 1  
! Set Supply address to 705  
! Set DMM address to 722  
! Print to the CRT  
190 Shunt_r= 1  
! Set value of current shunt  
200 !  
210 OUTPUT @Ps;"CLR"  
220 OUTPUT @Ps;"ID?"  
230 ENTER @Ps;Model$  
! Preset Supply to turn on state  
! Queries Supply model number  
! Enter model number  
240 !  
250 OUTPUT @Vm;"PRESET NORM;TRIG HOLD"  
! Preset DMM  
260 OUTPUT @Vm; “NPLC 100”  
! Set DMM sample  
270 !  
280 !  
290 !  
300 OUTPUT @Ps;"CMODE 1"  
310 !  
! Turn on Supply CAL mode  
320 IF Model$="Agilent6625A" OR Model$='Agilent6628A" THEN No_of_outputs=2  
330 IF Model$="Agilent6626A" OR Model$="Agilent6629A" THEN No_of_outputs=4  
340 !  
350 FOR Chan=1 TO No_of_outputs  
! Loop to test all outputs  
360  
370  
380  
390  
400  
410  
Get_data  
! Call program with range data  
! Call program to cal low range voltage  
! Call program to cal high range voltage  
! Call program to cal overvoltage  
! Call program to cal low range current  
! Call program to cal high range current  
! Test next output  
Cal_voltage(Vrng_lo)  
Cal_voltage(Vrng_hi)  
Cal_overvoltage(Vrng_hi)  
Cal_current(Irng_lo)  
Cal_current(Irng hi)  
420 NEXT Chan  
430 !  
440 Chan=l  
! Select channel to cal sink (-CC)  
! Select current source channel  
! Call program with range data  
! Call program to cal low range sink  
! 1 is + polarity for DMM  
450 Other_chan=2  
460 Get_data  
470 Cal_sink(Irng_lo,1)  
480  
490 Chan=2  
! Select channel to cal sink (-CC)  
! Select current source channel  
! Call program with range data  
! Call program to cal low range sink  
500 Other_chan=l  
510 Get_data  
520 Cal_sink(Irng_lo,-1)  
Alignment Procedures 95  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alignment Program (continued)  
530  
540  
550  
560 Chan=1  
570 Other_chan=2  
580 Get_data  
590 Cal_sink(Irng_hi,1)  
600 !  
! -1 is reverse polarity, this saves  
! reversing leads as sink channel  
! becomes the source for other channel  
! Select channel to cal sink (-CC)  
! Select current source channel  
! Call program with range data  
! Call program to cal high range sink  
610 Chan=2  
! Select channel to cal sink (-CC)  
! Select current source channel  
! Call program with range data  
! Call program to cal high range sink  
620 Other_chan=1  
630 Get_data  
640 Cal_sink(Irng_hi,-1)  
650 !  
660 IF No_of_outputs=4 THEN  
! Cal outputs 3 & 4  
670  
680  
690  
700  
710  
720  
730  
740  
750  
760  
770  
780  
790  
800  
810  
820  
Chan=3  
Other_chan=4  
Get_data  
Cal_sink(Irng_lo,1)  
!
Chan=4  
Other_chan=3  
Cal_sink(Irng_lo,-1)  
!
! Select channel to cal sink (-CC)  
! Select current source channel  
! Call program with range data  
! Call program to cal low range sink  
! Select channel to cal sink (-CC)  
! Select current source channel  
! Call program to cal low range sink  
Chan=3  
! Select channel to cal sink (-CC)  
! Select current source channel  
! Call program to cal high range sink  
Other_chan=4  
Cal_sink(Irng_hi,1)  
!
Chan=4  
Other_chan=3  
Cal_sink(Irng_hi,-1)  
! Select channel to cal sink (-CC)  
! Select current source channel  
! Call program to cal high range sink  
830 END IF  
840  
:
850 Check_error  
860 Cal_mode_off  
870 !  
! Call error checking program  
! Call program to turn cal mode off  
880 END  
890 !  
900 !  
910 !  
920 SUB Clear_screen  
! Subprogram to clear display  
! Clear screen  
930  
OUTPUT 2;CHR$(255)&CHR$(75);  
940 SUBEND  
950 !  
960 !  
970 !  
980 DEF FNDcv  
! Function to r at voltage  
990  
COM /Vm/ @Vm  
1000  
1010  
1020  
1030  
OUTPUT @Vm;"DCV"  
TRIGGER @Vm  
ENTER @Vm;Volts  
RETURN Volts  
! Set DMM to read DC voltage  
! Take a voltage reading  
! Enter the voltage  
! Return voltage reading  
! End of voltage function  
1040 FNEND  
96 Alignment Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alignment Program (continued)  
1050 !  
1060 !  
1070 !  
1080 DEF FNDci(I_range)  
! Function to read current  
1090  
1100  
1110  
1120  
1130  
1140  
1150  
1160  
1170  
1180  
COM /Vm/ @Vm  
COM /Shunt/ Shunt_r  
IF I_range=2 THEN  
Amps=FNDcv/Shunt_r  
ELSE!  
OUTPUT @Vm;"DCI"  
TRIGGER @Vm  
ENTER @Vm;Amps  
END IF  
! 2A must read across shunt  
! Convert shunt voltage to Amps  
! <2A read direct from DMM  
! Set DMM to read DC current  
! Take a current reading  
! Enter the current  
RETURN Amps  
! Return the current reading  
1190 FNEND  
1200 !  
1210 !  
1220 !  
1230 SUB Output_on  
! Subprogram to turn ON outputs  
! Turn on outputs 1 & 2  
1240  
1250  
1260  
1270  
1280  
COM /Ps/ Chan,Other_chan, @Ps, Model$[7]  
COM /Outputs/ No_of_outputs  
OUTPUT @Ps; OUT l,l;OUT 2,1  
IF No_of_outputs=4 THEN OUTPUT @Ps;"OUT 3,1;OUT 4,1"  
! Turn on outputs 3 & 4  
1290 SUBEND  
1300 !  
1310 !  
1320 !  
1330 SUB Output_off  
! Subprogram to turn OFF outputs  
! Turn off outputs 1 & 2  
1340  
1350  
1360  
1370  
1380  
COM /Ps / Chan, Other_chan, @Ps, Mode1$ [ 7 ]  
COM /Outputs/ No_of_outputs  
OUTPUT @Ps;"OUT 1,0,OUT 2,0"  
IF No_of_outputs=4 THEN OUTPUT @Ps; “OUT 3,0;OUT 4,0”  
! Turn off outputs 3 & 4  
1390 SUBEND  
1400 !  
1410 !  
1420 !  
1430 Cal_voltage : SUB Cal_voltage(V_range)  
! Subprogram to cal voltage  
1440  
1450  
1460  
1470  
1480  
1490  
1500  
1510  
1520  
1530  
1540  
1550  
1560  
COM /Ps/ Chan,Other_chan,@Ps,Model$ [7]  
!
IF V_range=50 THEN GOTO Cal_v  
Output_off  
! Skip setup instructions  
! Turn off all outputs  
Clear_screen  
PRINT "Set up output";Chan;"for voltage calibration  
PRINT  
PRINT " Use Fig A-1 of the Operating Manual"  
PRINT  
PRINT “ 3458A input is connected to VOLTAGE terminals”  
PRINT  
PRINT " The current shunt is NOT used"  
PRINT  
Alignment Procedures 97  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alignment Program (continued)  
PRINT "Press CONTINUE when ready  
1570  
1580  
1590  
1600  
1610  
1620  
1630  
1640  
1650  
1660  
1670  
1680  
1690  
1700  
1710  
1720  
1730  
1740  
1750  
1760  
1770  
1780  
1790  
1800  
1810  
1820  
1830  
1840  
1850  
1860  
1870  
1880  
1890  
1900  
1910  
PAUSE  
Output_on  
!
Cal_v:  
!
Clear_screen  
PRINT "Calibrating the";V_range;"volt range programming”  
!
OUTPUT @Ps;"VRSET" ;Chan,V_range  
!
OUTPUT @Ps; VLO";Chan  
WAIT 1  
Vlo=FNDcv  
!
OUTPUT @Ps; “VHI” ;Chan  
WAIT 1  
Vhi=FNDcv  
!
OUTPUT @Ps; “VDATA” ;Chan, Vlo, Vhi  
!
Clear_screen  
PRINT "Calibrating the";V_range,'volt range readback"  
!
OUTPUT @Ps; VRLO";Chan  
REPEAT  
UNTIL BIT(SPOLL(@Ps),4)  
Vrlo=FNDcv  
!
OUTPUT @Ps; “VRHI”;Chan  
REPEAT  
UNTIL BIT(SPOLL(@Ps),4)  
Vrhi=FNDcv  
!
OUTPUT @Ps; “VRDAT”; Chan, Vrlo, Vrhi  
!
! Set voltage range  
! Set to low output voltage  
! Wait for supply to stabilize  
! Read low voltage  
! Set to high output voltage  
! Wait for supply to stabilize  
! Read high voltage  
! Send output voltage data  
! Set low readback voltage  
! Wait for supply to finish  
! Finished when Bit 4 goes true  
! Read low output voltage  
! Set high readback voltage  
! Wait for supply to finish  
! Finished when Bit 4 goes true  
! Read high output voltage  
! Send voltage readback data  
1920 SUBEND  
1930 !  
1940 !  
1950 !  
1960 Cal_overvoltage:SUB Cal_overvoltage(V_range)  
! Sub program to cal OV  
1970  
1980  
1990  
2000  
2010  
2020  
2030  
2040  
2050  
2060  
COM /Ps/ Chan,Other_chan,@Ps,Model$[7]  
!
Clear_screen  
PRINT "Calibrating the Overvoltage circuit  
!
OUTPUT @Ps;"VRSET";Chan,V_range  
OUTPUT @Ps;"OVCAL";Chan  
REPEAT  
UNTIL BIT(SPOLL(@Ps),4)  
!
! Set voltage range  
! Calibrate Overvoltage  
! Wait for supply to finish  
! Finished when Bit 4 goes true  
2070 SUBEND  
2080 !  
98 Alignment Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alignment Program (continued)  
2090 !  
2100 !  
2110 Cal_current: SUB Cal_current(I_range)  
! Subprogram to cal current  
! Skip setup instructions  
2120  
2130  
2140  
2150  
2160  
2170  
2180  
2190  
2200  
2210  
2220  
2230  
2240  
2250  
2260  
2270  
COM /Ps/ Chan,Other_chan,@Ps,Model$[7]  
!
IF I_range= .5 THEN GOTO Cal_i  
!
Output_off  
!
Clear_screen  
PRINT "Set up output";Chan;"for";I_range;”A current calibration”  
PRINT  
IF I_range=2 THEN  
! If the next cal range is 2  
! amps then print this message  
PRINT " Use Fig A-3 of the Operating Manual"  
PRINT  
PRINT “3458A input is connected to VOLTAGE terminals”  
PRINT  
PRINT “The current shunt IS used”  
2280 ELSE  
! If not 2 amps then print this  
2290  
2300  
2310  
2320  
2330  
PRINT “Use Fig A-2 of the Operating Manual”  
PRINT  
PRINT “3458A input is connected to CURRENT terminals”  
PRINT  
PRINT “The current shunt is NOT used”  
2340 END IF  
2350 PRINT  
2360 PRINT “Press CONTINUE when ready”  
2370  
2380  
2390  
PAUSE  
Output_on  
!
2400 Cal_i :  
!
2410  
2420  
2430  
2440  
2450  
2460  
2470  
2480  
2490  
2500  
2510  
2520  
2530  
2540  
2550  
2560  
2570  
2580  
2590  
2600  
Clear_screen  
PRINT "Calibrating the";I_range;"A current programming  
!
OUTPUT @Ps;"IRSET";Chan,I_range  
OUTPUT @Ps;''ILO'';Chan  
WAIT 1  
Ilo=FNDci(I_range)  
!
OUTPUT @Ps;"IHI";Chan  
WAIT 1  
Ihi=FNDci(I_range)  
!
OUTPUT @Ps;”IDATA”;chan,Ilo,Ihi  
!
! Set current range  
! Set output current low  
! Wait for supply to stabilize  
! Read low output current  
! Set output current high  
! Wait for supply to stabilize  
! Read high output current  
! Set output current data  
Clear_screen  
PRINT 'Calibrating the";I_range;"A current readback"  
!
OUTPUT @Ps;"IRLO";Chan  
REPEAT  
! Set low readback current  
! Wait for supply to finish  
! Finished when Bit 4 goes true  
UNTIL BIT(SPOLL(@Ps),4)  
Alignment Procedures 99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alignment Program (continued)  
2610  
2620  
2630  
2640  
2650  
2660  
2670  
2680  
2690  
Irlo=FNDci(I_range)  
!
OUTPUT @Ps;"IRHI";Chan  
REPEAT  
UNTIL BIT(SPOLL(@Ps),4)  
Irhi=FNDci(I_range)  
!
! Read low output current  
! Set high readback current  
! Wait for supply to finish  
! Finished when Bit 4 goes true  
! Read high output current  
OUTPUT @Ps;"IRDAT";Chan,Irlo,Irhi  
!
! Send current readback data  
2700 SUBEND  
2710 !  
2720 !  
2730 !  
2740 Cal_sink SUB Cal_sink(I_range,Polarity)  
! Subprogram to cal -CC  
! Skip setup instructions  
2750  
2760  
2770  
2780  
2790  
2800  
2810  
2820  
2830  
2840  
2850  
2860  
2870  
2880  
2890  
2900  
2910  
2920  
2930  
2940  
2950  
2960  
2970  
2980  
2990  
3000  
3010  
3020  
3030  
3040  
3050  
3060  
3070  
3080  
3090  
3100  
COM /Ps/ Chan,Other_chan,@Ps,Model$[7]  
!
IF I_range= .5 THEN GOTO Sink_rb  
IF Model$="Agilent6625A" AND I_range=2 THEN I_range= .5  
! If a 6625A, 2A output will be  
! set to 5A to match output 1  
! Skip setup instructions  
IF Chan=2 OR Chan=4 THEN GOTO Sink_rb  
!
Output_off  
Clear_screen  
PRINT “Set up output';Chan;”for ;I_range; “A current SINK cal”  
PRINT  
PRINT “ Connect outputs”;Chan; “and” ;Other_chan; “together as in”  
!
IF I range=2 THEN  
PRINT " Fig A-5 of the Operating Manual'  
PRINT  
PRINT " 3458A input is connected to VOLTAGE terminals"  
PRINT  
PRINT " The current shunt IS used for this part"  
PRINT " of the alignment"  
ELSE  
PRINT “Fig A-4 of the Operating Manual”  
PRINT  
PRINT “3458A input is connected to CURRENT terminals”  
PRINT  
! If it is 2A output use setup 5  
! If not 6626A 2A use setup 4  
PRINT " The current shunt IS NOT used"  
END IF  
PRINT  
PRINT “The sink readback of both outputs”  
PRINT “will be aligned with this setup”  
!
PRINT  
PRINT "Press CONTINUE when ready”  
PAUSE  
!
3110 Sink_rb  
3120  
Clear_screen  
100 Alignment Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alignment Program (continued)  
PRINT "Calibrating output";Chan;"";I_range;"A current sink readback"  
3130  
3140  
3150  
3160  
3170  
3180  
3190  
3200  
3210  
3220  
3230  
3240  
3250  
3260  
3270  
3280  
3290  
3300  
3310  
3320  
3330  
3340  
3350  
3360  
3370  
!
OUTPUT @Ps;"IRSET";Chan,I_range;";IRSET";Other chan,I_range  
! Set I range for source and  
! sink range  
OUTPUT @Ps; “ISET”;Other_chan,0;”;VSET”;Other_chan,7  
! Set source V & I output  
Output_on  
WAIT 1  
!
OUTPUT @Ps;"IRLN";Chan  
REPEAT  
UNTIL BIT(SPOLL(@Ps),4)  
Ilo=(FNDci(I_range))*Polarity  
! Wait for source to stabilize  
! Set sink readback low  
! Wait for supply to finish  
! Finished when Bit 4 goes true  
! Read low sink current  
!
OUTPUT @Ps;"ISET";Other_chan,I_range  
WAIT 1  
!
OUTPUT @Ps;"IRHN";Chan  
REPEAT  
UNTIL BIT(SPOLL(@Ps),4)  
Ihi=(FNDci(I_range))*Polarity  
! Set source output current  
! Set sink readback high  
! Wait for supply to finish  
! Finished when Bit 4 goes true  
! Read high sink current  
!
OUTPUT @Ps; NIDAT”;chan,Ilo,Ihi  
!
! Send current sink data  
3380 SUBEND  
3390 !  
3400 !  
3410 !  
3420 Get_data SUB Get_data  
! Subprogram with range data  
3430  
3440  
3450  
3460  
3470  
3480  
3490  
3500  
3510  
3520  
3530  
3540  
3550  
3560  
3570  
3580  
3590  
COM /Specs/ Vrng_lo,Vrng_hi,Irng_lo,Irng_hi  
COM /Ps/ Chan,Other_chan,@Ps,Model$[7]  
!
IF Model$="Agilent6625A" AND Chan=1 THEN RESTORE Type_1_data  
IF Model$="Agilent6625A" AND Chan=2 THEN RESTORE Type_2_data  
IF Model$=”Agilent6626A" AND Chan=1 THEN RESTORE Type_1_data  
IF Model$="Agilent6626A" AND Chan=2 THEN RESTORE Type 1 data  
IF Model$=“Agilent6626A" AND Chan=3 THEN RESTORE Type_2_data  
IF Model$=“Agilent6626A" AND Chan=4 THEN RESTORE Type_2_data  
IF Model$=“Agilent6628A" OR Model$=“Agilent6629A" THEN RESTORE Type_2_data  
!
READ Vrng_lo,Vrng_hi,Irng_lo,Irng_hi  
!
!
! Read the Range data  
!
!
!
Output variables  
Vrng  
range data  
Irng  
Lo Hi  
.015, .500  
Lo Hi  
3600 Type 1 data:DATA  
3610 Type_2_data:DATA  
3620 !  
7, 50,  
16, 50,  
! Type 1 data (25W output)  
! Type 2 data (50W output)  
.200,  
2
3630 SUBEND  
3640 !  
Alignment Procedures 101  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alignment Program (continued)  
3650 !  
3660 !  
3670 Check_error:SUB Check_error  
! Subprogram to check for errors  
3680  
3690  
3700  
3710  
3720  
3730  
3740  
3750  
3760  
3770  
COM /Ps/ Chan,Other_chan,@Ps,Model$[7]  
Clear_screen  
PRINT “Checking for errors”  
OUTPUT @Ps;”ERR?”  
ENTER @Ps;Err  
IF Err>0 THEN  
PRINT “Error”;Err;”occurred. Supply may not be calibrated”  
PRINT "The alignment program is stopped"  
STOP  
! Query supply for errors  
! Enter the error number  
! If an error then print msg  
END IF  
3780 SUBEND  
3790 !  
3800 !  
3810 !  
3820 Cal_mode_off:SUB Cal_mode_off  
! Subprogram to turn off cal mod  
3830  
3840  
3850  
3860  
3870  
3880  
3890  
COM /Ps/ Chan,Other_chan,@Ps,Model$[7]  
Clear_screen  
PRINT "wait while calibration constants are stored"  
OUTPUT @Ps;”CMODE 0”  
OUTPUT @Ps;”CLR”  
! Turn cal mode off, store data  
! Clear supply  
Clear_screen  
PRINT "Calibration complete"  
3900 SUBEND  
3910 !  
102 Alignment Procedures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
B
Programming With a Series 200/300 Computer  
Introduction  
The purpose of this appendix is to serve as an introduction to programming your power supply with an HP Series 200/300  
computer using the BASIC language. Examples are included that employ some of the most frequently used functions.  
These examples have been written so that they will run on all power supplies. The values used in the examples (5 V and 0.5  
A for instance), are within the operating locus of all outputs on all models. The examples program only channels one and  
two because both models contain at least two channels (outputs).  
You must be familiar with the BASIC language to understand the examples. If you do not recognize a programming  
statement, look up the keyword in the BASIC Language Reference document that was supplied with your computer, and  
look up the device command in Chapter 5 of this manual. Questions about program structure and selection are discussed in  
the BASIC Programming Techniques manual.  
I/O Path Names  
Throughout this appendix, I/O path names are used in place of interface and device select codes. In a large program, I/O  
path names simplify changing the address of an instrument if necessary. Reading and writing the program is easier as well.  
The I/O path name can be carried in a common block and changed by a single assign statement.  
In the programming examples in this appendix, the I/O path name @Ps is used for the power supply. The ASSIGN  
statement that defines the I/O path must precede any statements that use the I/O path name. Therefore, instead of using the  
statement OUTPUT 705;VSET1,5in the following programs, the equivalent OUTPUT @Ps;"VSET1,5" statement is  
used. The examples assume that the power supply is at address 5 and the GP-IB interface in the computer is at select code  
7(factory default).  
Voltage and Current Programming  
The power supply normally functions in one of two modes, either constant voltage with current limit or constant current  
with voltage limit. The operating mode is determined by a combination of voltage and current settings and load resistance.  
For example, with a 50 load connected to output 1, the following program will put output 1 in constant voltage mode at 5  
volts out with a 0.5 amp current limit. In this case the output current would be 0.1 A.  
10 ASSIGN @Ps TO 705  
20 OUTPUT @Ps;VSET1,5;ISET1,0.5’  
30 END  
Line 10: Assigns the I/O pathname to the power supply.  
Line 20: Sets output voltage and current. Note the use of the semicolon to separate multiple device commands.  
If a 4 load were used instead of a 50 load, output 1 would have been operating in constant current mode at 0.5 amp out  
with a voltage limit of 5 volts. In this case the output voltage would be 2 V.  
Programming with a Series 200/300 Computer 103  
Download from Www.Somanuals.com. All Manuals Search And Download.  
VOLTAGE AND CURRENT PROGRAMMING WITH VARIABLES  
You can use variables in a program to represent data values in the device commands. This is useful in applications that  
require changing the voltage and current values to different predetermined settings. The following program uses a variable  
in a FOR NEXT loop to ramp up output voltage in 0.1 volt steps from 0 to 5 volts.  
10 ASSIGN @Ps TO 705  
20 OUTPUT @Ps;"CLR;ISET1,0.5"  
30 FOR Voltage=0 TO 5 STEP 0.1  
40 OUTPUT @Ps;VSET1,";Voltage  
50 WAIT 0.2  
60 NEXT Voltage  
70 END  
Line 10:  
Line 20:  
Line 30,60:  
Line 40:  
Assigns the l/O pathname to the power supply.  
Initializes the power supply to its power on state and sets the current limit.  
Increments the voltage in 0.1 V steps to 5 volts.  
Sets the voltage of output 1 to the value of the variable Voltage’. The comma inside the quotes is  
required because it separates numbers in the device command (the output channel number from the  
voltage value in this case). A space < SP > may also be used instead of the comma. The semicolon  
outside the quotes is used because it suppresses the <CR> < LF> that the computer would normally send  
to the power supply if a comma were used as a separator after a string item. Using a comma in this case  
would produce a syntax error in the power supply.  
Line 50:  
Waits 0.2 seconds between steps.  
Another way to use variables to represent data values in device commands is when using input statements to program the  
power supply. The following program uses input statements to program the voltage and current settings of output 1 and  
output 2.  
10 ASSIGN @Ps TO 705  
20 INPUT "ENTER A VOLTAGE FOR OUTPUT #1",V1  
30 INPUT ENTER A CURRENT LIMIT FOR OUTPUT #1",I1  
40 INPUT "ENTER A VOLTAGE FOR OUTPUT #2",V2  
50 INPUT "ENTER A CURRENT LIMIT FOR OUTPUT #2",I2  
60 OUTPUT @Ps;"VSET1,";V1;";ISET1,;I1;;VSET2,";V2;";ISET2,;I2  
70 END  
Line 10:  
Assigns the I/O pathname to the power supply.  
Line 20,30:  
Line 40,50:  
Line 60:  
Enters voltage and current values for output 1.  
Enters voltage and current values for output 2.  
Sets the voltage and current of outputs 1 and 2 to the values entered into the variables. The previous  
example explained the use of the comma inside the quotes and the semicolon that precedes the variable.  
The semicolon that follows the variable suppresses the comma that the computer would normally send to  
the power supply if a comma were used as a separator after a numeric item. The leading semicolons  
inside the quotes separate multiple device commands (the VSET commands from the ISET commands in  
this case).  
Voltage and Current Readback  
Reading back data from the power supply requires two statements. First, an output statement is used to query the power  
supply. A list of queries appears in Table 5-2 on page 70. The power supply responds to the query by entering the requested  
data into a buffer. Next, an enter statement is used to read the data from the buffer on the power supply into a variable in the  
computer. The following program queries the power supply for the voltage and current settings of output 1 and prints the  
results on the screen.  
104 Programming with a Series 200/300 Computer  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10 ASSIGN @Ps TO 705  
20 OUTPUT @Ps;VSET?1’  
30 ENTER @Ps;Vsl  
40 OUTPUT @Ps;"ISET?1"  
50 ENTER @Ps;Isl  
60 PRINT VOLTAGE SETTING OF OUTPUT #1 = ";Vsl  
70 PRINT "CURRENT LIMIT SETTING OF OUTPUT #1 = ";Isl  
80 END  
Line 10:  
Assigns the I/O pathname to the power supply.  
Line 20,30:  
Queries the supply for output 1s voltage setting. You cannot string multiple queries together in a single  
device command because the power supply can only return the most recently queried data.  
Queries the supply for output 1s current limit.  
Line 40,50:  
Line 60,70:  
Prints the results of the queries on the screen.  
Programming Power Supply Registers  
Present Status  
The power supply makes available several forms of status information. Chapter 5 discusses the different registers and their  
functions. The present status register contains continuously updated information. The following example shows how to  
query bit position 0 of output 1s status register to see if output 1 is in CV mode. In this example, the program references the  
function as a variable in a conditional execution statement. Note that instead of printing a message, line 30 could be used to  
branch to another part of the program in the event that the supply is in CV mode.  
10 ASSIGN @Ps TO 705  
20 COM /Ps/ @Ps  
30 IF FNCv _ mode THEN  
40 PRINT "OUTPUT 1 IS IN CV MODE"  
50 END IF  
60 END  
70 !  
80 !  
90 DEF FNCv _ mode  
100 COM /Ps/ @Ps  
110 OUTPUT @Ps;STS?1"  
120 ENTER @Ps;Sts  
130 RETURN BIT(Sts,0)  
140 FNEND  
Line 10:  
Line 20:  
Assigns the I/O path name to the power supply.  
Declares a common block for the I/O path name. The COM statement must be used for the @Ps variable  
to preserve its value in the function subprogram.  
Line 30,40,50: If FNCv _ mode is true, print the message.  
Line 90:  
Defines the Cv _ mode function.  
Line 100:  
Line 110,120:  
Line 130:  
Brings in the common block for the I/O pathname.  
Reads the present status of output 1 into the variable Sts.  
Returns the value of bit position 0 of Sts.  
Programming with a Series 200/300 Computer 105  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Service Request and Serial Poll  
The fault and mask registers, when used in conjunction with the service request and serial poll functions, allow you to  
select which conditions can cause computer interrupts. The fault and mask registers can also be used independently of the  
serial poll or service request if so desired. The following example shows how to enable an interrupt to the computer in the  
case of an overvoltage condition. After the interrupt has occurred, this example includes an interrupt routine that conducts a  
serial poll to determine on which output the overvoltage occurred. Note that this example assumes that terminal block  
external OV trip lines are not wired together.  
10 ASSIGN @Ps TO 705  
20 COM /Ps/ @Ps  
30 OUTPUT @Ps;CLR;UNMASK1,8;UNMASK2,8;SRQ1’  
40 ON INTR 7,1 CALL Err _ trap  
50 ENABLE INTR 7;2  
60 OUTPUT @Ps;"OVSET1,4;OVSET2,4"  
70 OUTPUT @Ps;VSET1,5;VSET2,5"  
80 Lbl: GOTO Lbl  
90 END  
100 !  
110 !  
120 SUB Err _ trap  
130 OFF INTR  
140 COM /Ps/ @Ps  
150 IF BIT(SPOLL(@Ps),0) THEN  
160 OUTPUT @Ps;OUT1,0;OVRST1"  
170 PRINT "OVERVOLTAGE ON OUTPUT #1"  
180 END IF  
190 IF BIT(SPOLL(@Ps),1) THEN  
200 OUTPUT @Ps;"OUT2,0;OVRST2’  
210 PRINT OVERVOLTAGE ON OUTPUT #2"  
220 END IF  
230 OUTPUT @Ps;FAULT?1;FAULT?2’  
240 SUBEND  
LINE 10:  
LINE 20:  
Assigns the I/O pathname to the power supply.  
Declares a common block for the I/O pathname. The COM statement must be used for the @Ps variable  
to preserve its value for use in the service routine.  
LINE 30:  
Returns the power supply to its power on state, unmasks output 1s and output 2s OV status bits to  
generate faults, and enables the service request capability on all outputs.  
Defines interrupt at interface 7, with GP-IB priority 1.  
Enables interrupt at interface 7 for service request type interrupts only.  
Sets the overvoltage of outputs 1 and 2 to 4 volts.  
LINE 40:  
LINE 50:  
LINE 60:  
LINE 70:  
Sets the voltage of outputs 1 and 2 above the OV setting so that both outputs will overvoltage when the  
program is run.  
LINE 80:  
Waits for the computer to receive the interrupt. This simulates conditions that would normally exist when  
a program is running.  
LINE 120:  
LINE 130:  
LINE 140:  
Defines the error handling routine.  
Disables interrupt capability while processing.  
Brings in the common block for the I/O pathname.  
LINE 150-180: Conducts a serial poll. If bit 0 in serial poll register indicates a fault for output 1, output 1 is disabled and  
the overvoltage circuit is reset.  
LINE 190-220: Checks bit 1 in serial poll register for a fault on output 2. If true, output 2 is disabled and the overvoltage  
circuit is reset.  
LINE 230:  
Reads fault registers to clear FAU bits in serial poll register.  
106 Programming with a Series 200/300 Computer  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Error Detection  
The power supply can recognize programming errors and can inform you when a programming error occurs. When an error  
is detected, no attempt is made to execute the command. Instead, a bit in the serial poll register is set. If SRQ2 or SRQ3 is  
set, an interrupt will be generated. The following program checks for programming errors and can be entered and run as is.  
While it is running, commands can be sent to the power supply from the keyboard. If the ERROR annunciator on the power  
supplys front panel indicates that an error has been detected, depress the labeled softkey to display the error on your  
computer screen.  
10 ASSIGN @Ps TO 705  
20 COM /Ps/ @Ps  
30 ON KEY 0 LABEL "ERROR?CALL Err_trap  
40 Lbl: GOTO Lbl  
50 END  
60 !  
70 !  
80 SUB Err _ trap  
90 OFF KEY  
100 COM /Ps/ @Ps  
110 OUTPUT @Ps;"ERR?’  
120 ENTER @Ps:Err  
130 OUTPUT 2 USING "#,K";CHR$(255)&CHR$(75)  
140 IF Err THEN  
150 PRINT "POWER SUPPLY PROGRAMMING ERROR:’  
160 END IF  
170 SELECT Err  
180 CASE 0  
190 PRINT NO ERROR OCCURRED"  
200 SUBEXIT  
210 CASE 1  
220 PRINT "INVALID CHARACTER’  
230 CASE 2  
240 PRINT "INVALID NUMBER"  
250 CASE 3  
260 PRINT "INVALID STRING"  
270 CASE 4  
280 PRINT SYNTAX ERROR"  
290 CASE 5  
300 PRINT NUMBER OUT OF RANGE"  
310 CASE 6  
320 PRINT "DATA REQUESTED WITHOUT QUERY’  
330 CASE 7  
340 PRINT STRING EXCEEDS DISPLAY LENGTH"  
350 CASE 8  
360 PRINT "NUMBER TOO LARGE FOR INPUT BUFFER"  
370 CASE 28  
380 PRINT "INVALID CHARACTERS IN STRING"  
390 CASE ELSE  
400 PRINT "UNRECOGNIZED ERROR NUMBER’; Err  
410 END SELECT  
420 PRINT "RE-ENTER STATEMENT AND TRY AGAIN"  
430 SUBEND  
Programming with a Series 200/300 Computer 107  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LINE 10:  
LINE 20:  
LINE 30:  
LINE 40:  
LINE 80:  
LINE 90:  
LINE 100:  
Assigns the I/O path name to the power supply.  
Declares a common block for the I/O path name.  
Defines interrupt on softkey depression and branch to error routine.  
Idles on softkey definition.  
Defines subprogram Err _ trap  
Disables interrupt capability while processing.  
Brings in the common block for the I/O pathname.  
LINE 110,120: Enters error code from power supply.  
LINE 130: Clears computer screen.  
LINE 140,150: If an error occurred, print message.  
LINE 170-420: Prints message based on error code. Prompt user to try again. If no error occurred, print message saying  
no error occurred.  
Stored Operating States  
Your power supply has registers that can store up to 11 operating states (note registers 0-3 are non-volatile, see Chapter 3).  
These states can be recalled in arbitrary order. Storing a state involves taking a "snapshot" of the voltage and current  
settings that are in effect when the command is received. The following example uses stored operating states to set up an  
output. Using this method of setting up outputs saves processing time and facilitates repeating the same commands.  
10 ASSIGN @Ps TO 705  
20 OUTPUT @Ps;OUT1,0;OUT2,0"  
30!  
40 OUTPUT @Ps;"VSET1,1;ISET1,0.01;VSET2,5;ISET2,0.01;STO1"  
50 OUTPUT @Ps;"VSETl,2;ISETl,0.02;VSET2,4;ISET2,0.02;STO2"  
60 OUTPUT @Ps;"VSETl,3;ISETl,0.03;VSET2,3;ISET2,0.03;STO3"  
70 OUTPUT @Ps;"VSETl,4;ISETl,0.04;VSET2,2;ISET2,0.04;STO4"  
80 OUTPUT @Ps;VSET1,5;ISET1,0.05;VSET2,1;ISET2,0.05;STO5"  
90 !  
100 OUTPUT @Ps;"CLR"  
110 FOR State = 1 to 5  
120 OUTPUT @Ps;"RCL";State  
130 WAIT 2  
140 NEXT State  
150 END  
LINE 10:  
Assigns the I/O pathname to the power supply.  
LINE 20:  
Disables Output 1 and Output 2.  
LINE 40-80:  
Stores 5 operating states for Output 1 and Output 2 in storage registers 1 through 5. Outputs not explicitly  
programmed will store the settings that are in effect when the store command is received.  
Clears the supply. All outputs are enabled and set the initial power on state.  
LINE 100:  
LINE 110-140: Loops through the sequence of five states with a two second wait between states.  
108 Programming with a Series 200/300 Computer  
Download from Www.Somanuals.com. All Manuals Search And Download.  
C
Command Summary  
Introduction  
For convenience, a “quick reference” listing tabulating general command headers, and calibration command headers is  
shown at the beginning of Table C-1 (page 110). (Note that calibration commands are described in detail in Appendix A.)  
Table C-2 (pages 111-115) provides an alphabetical listing and a brief description of each command that may be sent to the  
power supplies. Chapter 5 gives a complete description (including syntax) of most of the commands listed.  
All of the commands can be executed remotely over the GP-IB. Many of the commands may also be executed locally from  
the supply's front panel (marked by single asterisks in Table C-1).  
Several commands will generate a response which will be indicated on the front panel of the supply. These commands are  
indicated by double asterisks in Table C-1.  
Command headers are accepted in upper or lower case letters although only upper case letters are used in this summary.  
The brackets < > indicate a number to be sent.  
Note that <ch> must specify an output channel number from 1 through 4 as applicable. Commas or spaces are required  
between numbers. Spaces are optional between the command header and the first number. Use semicolons between  
multiple commands sent in one statement.  
Command Summary 109  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table C-1 Quick Reference Listing of Commands  
General Commands  
ASTS? < ch >  
CLR  
OUT < ch >, < 1 (on) or 0 (off) >  
OUT? < ch >  
OVRST < ch >  
OVSET < ch > < x >  
OVSET? < ch >  
PON < 1 (on) or 0 (off) >  
PON?  
RCL < register >  
ROM?  
SROM? < ch >  
*
**  
*
CMODE < 1 (on) or 0 (off) >  
CMODE?  
DCPON < 1 (on) or 0 (off) >  
DLY < ch >, < delay time >  
DLY? < ch >  
DSP < 1 (on) or 0 (off)>  
DSP <"string">  
DSP?  
**  
*
ERR?  
SRQ <x>  
*
FAULT? < ch >  
ID?  
IOUT? <ch>  
SRQ?  
STO < register >  
STS? < ch >  
**  
IRSET <ch>, <x>  
IRSET? < ch >  
ISET < ch >, < current setting >  
ISET? <ch>  
ISTEP < ch >, < step value >  
METER < ch >  
METER?  
OCP < ch >, < 1, (on) or 0 (off) >  
OCP? < ch >  
TEST?  
*
**  
*
UNMASK < ch >, < sum of bit weights >  
UNMASK? < ch >  
VOUT? < ch >  
VRSET < ch >, <x >  
VRSET? < ch >  
VSET < ch >, < voltage setting >  
VSET? < ch >  
VSTEP < ch >, < step vaiue >  
**  
**  
**  
**  
**  
**  
**  
*
OCRST <ch>  
Calibration Commands  
NIDAT < ch >, < Ilo >, < Ihi >  
IDATA < ch >, < llo >, < ihi >  
IHI < ch >  
OVCAL <ch>  
ILO < ch >  
IRDAT < ch > , < Ilo >, < Ihi >  
IRHI < ch >  
VDATA < ch >, < VLo >, < Vhi >  
VHI < ch >  
VLO < ch >  
IRLO < ch >  
IRLN < ch >  
VRDAT < ch >, < Vlo >, < Vhi >  
VRHI < ch >  
IRHN < ch >  
VRLO < ch >  
Servlce Commands  
IDAC < ch >, < x >  
IDAC? < ch >  
VDAC < ch >,< x >  
VDAC? < ch >  
RDAC < ch >,< x >  
RDAC? < ch >  
VMUX? < ch >  
*· = Command can be executed from the front panel.  
** = Data returned from these cormnands is also displayed on the front panel after command is issued. Note that the display  
must be enabled (i.e. DSP = 1).  
110 Command Summary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table C-2 Command Summary  
ASTS? < ch >  
Queries the accumulated status (ASTS) of the specified output < ch > . The response  
(integer 0-255) represents the sum of the binary weights of the ASTS register bits (see  
page 77). The ASTS register is automatically set to the present status after being  
queried.  
CLR  
Returns the entire power supply (all outputs) to the power on state, except that the  
supply is not unaddressed and its store/recall registers are not changed (see page 76).  
CMODE < on/off >  
CMODE?  
Turns the calibration mode on or off. On/off is a 1 to turn the calibration mode on; a 0  
turns it off (see Appendix A).  
Queries if the calibration mode is on or off. Response is either a 1 (on) or 0 (off), (see  
page 82).  
DCPON < on/off >  
Sets the state of the outputs at power-on. For on/off = 0, all outputs will be off when the  
power supply is turned on. For on/off = 1 all outputs will be on when the power supply  
is turned on.  
DLY < ch >,< x >  
Sets the reprogramming delay time (x) for the specified output < ch > . This delay is  
used to mask the CV, + CC, - CC, and UNR status bits from the fault register and the  
OCP function for the specified delay period (0 to 32 seconds). The delay time is initiated  
after a VSET, ISET, OUT, RCL, over voltage reset, or over current reset commands is  
sent (see page 81).  
DLY? < ch >  
DSP < on/off>  
DSP?  
Queries the present reprogramming delay of the specified output <ch> (see page 81).  
The response can be a real number from 0.000 to 32 seconds (e.g. 0.450 seconds).  
Turns the power supplys front panel display either on or off (see page 81). On/off equals  
1 to turn the display on or a 0 to turn it off.  
Queries the present status of the display (see page 81). Response is either a 1 (on), or a 0  
(off).  
DSP “xxxxxxxxxxxx”  
Puts the quoted string on the power supply's front panel display (see page 81). Only  
numerals, upper case letters, and spaces are allowed (12 characters max) in the quoted  
string.  
ERR?  
Queries the present programming or hardware error (see page 82). An error code number  
is returned over the GP-IB to identify the error. In the local mode, pressing the error key  
will cause the appropriate error message (not the error code) to be displayed on the front  
panel. The error register is cleared after being read.  
FAULT? < ch >  
Queries the fault register of the specified output < ch > (see page 77) . A bit is set in the  
fault register when the corresponding bit is set in both the status and the mask registers.  
The response is an integer 0 to 255. The fault register is cleared after being read.  
IDAC < ch >, < x >  
IDAC? < ch >  
Programs the current DAC to a specified number of counts < x > for the selected  
channel < ch > . The number returned is in DAC counts.  
Queries the LSB setting of the current DAC for the selected channel < ch > . The  
number returned is in DAC counts.  
Command Summary 111  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table C-2 Command Summary (continued)  
ID?  
Queries the identification (model number) of the supply. (see page 82).  
IDATA < ch >, < Ilo > , < Ihi > Sends data to calibrate the current setting circuits of the specified output < ch >. Ilo and  
Ihi are measured values which the supply uses to calculate correction constants (see  
Appendix A).  
IHI < ch >  
Causes the current of the specified output < ch > to go to the high calibration point (see  
Appendix A).  
ILO < ch >  
IOUT? < ch >  
Causes the current of the specified output < ch > to go to the low calibration point (see  
Appendix A).  
Queries the measured output current of the specified output < ch > (see page 73). The  
response is a real number. The front panel display can be used to monitor the measured  
output current (and voltage) of the selected output < ch >.  
IRDAT < ch >, < Ilo >, < Ihi >  
IRLO < ch >  
Used to calibrate the + current readback for the specified output < ch >. This command  
must be sent after IRHI and IRLO commands.  
Finds the minimum + current readback value for the specified output < ch > ( see  
Appendix A.)  
IRHI < ch >  
Funds the maximum + current readback value for the specified output < ch > (see  
Appendix A.)  
IRLN <ch>  
Finds the minimum - current readback value for the specified output < ch > (see  
Appendix A.)  
IRHN <ch>  
Finds the maximum - current readback value for the specified output < ch > ( see  
Appendix A).  
IRSET < ch >, < x >  
Sets the full scale current range programming for an output < ch > . The power supply  
will select the lowest operating range that will contain the value (x). If the present  
operating current of the supply is greater than the full scale range selected, the present  
output current will be reduced to the full scale value of the new range selected. If the  
current entered is larger than the highest possible range, an error message will result.  
IRSET? < ch >  
ISET < ch >,< x >  
ISET? < ch >  
Queries the full scale current range setting of an output < ch > .  
Sets the current of the specified output < ch > (see page 73).  
Queries the present current setting of the specified output < ch > (see page 73). The  
response is a real number.  
ISTEP < ch >,< x >  
*
Increases or decreases an output < ch > current in steps. If the step value < x > is  
positive, output current will increase by the step value, if the step value is negative, the  
output current will decrease by the step value. Trying to step beyond the active operating  
range will cause a number range error.  
METER < ch >  
METER?  
Selects which output < ch > of the supply the front panel will display.  
Returns which output the supply is displaying on the front panel.  
112 Command Summary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table C-2 Command Summary (continued)  
NIDAT < ch > . < Ilo > . < Ihi > Sends data to calibrate - current readback for the selected output < ch > . This command  
must be sent after IRLN and IRHN commands.  
OCP < ch >, < x >  
Enables the over current protection circuit protection circuit for the specified output  
< ch >. This circuit, when enabled, causes the output to go to the off state when the  
output is on the + CC mode. On/off is a 1 to turn on (enable) or a 0 to turn off (disable)  
the circuit. (see page 75).  
OCP? < ch >  
Queries the over current protection circuit on/off status for the specified output < ch >  
(see page 75). Response is either a 1 (on) or a 0 (off). The OCP ENBLD annunciator on  
the front panel displays the on/off status of the OCP circuit for the selected output.  
OCRST < ch >  
Returns the specified output < ch > to the previous settings after it had been turned off  
by the over current protection circuit (see page 75).  
OUT < ch >, < on/off >  
OUT? < ch >  
Turns the specified output < ch > on or off. On/off equals 1 to turn the output on and  
equals 0 to turn the output off (see page 77).  
Queries whether the specified output < ch > is turned on or off (see page 77). The  
response is either 1 (on) or 0 (off). The front panel will display the message  
’’DISABLEDwhen the selected output channel is turned off.  
OVCAL < ch >  
OVRST < ch >  
Causes the specified output < ch > to go through the overvoltage calibration routine (see  
Appendix A).  
Attempts to reset the overvoltage crowbar circuit in the specified output < ch > (see page  
77).  
OVSET <ch>,<x>  
OVSET? < ch >  
Sets the overvoltage trip point < x > for the specified output < ch > ( see page 77).  
Queries the present overvoltage setting of the specified output < ch > (see page 77). The  
response is a real number.  
PON < on/off >  
Enables the power on service request (SRQ). On/off = 1 causes the power supply to  
generate a service request when power is applied (see page 80). On/off = 0 disables the  
PON SRQ. The on/off settings is retained in the supplys memory through interruption of  
ac line power.  
PON?  
Queries the present state of the power on SRQ function (see page 80). The response is  
either 1 (on) or 0 (off).  
RCL < x >  
Recalls the voltage and current settings for all output channels from the specified  
internal register < x >. These settings were previously stored using the STO command.  
RCL programs all output channels in sequential order (1, 2, 3, 4) to these stored settings  
(see page 80).  
RDAC < ch >, < counts >  
RDAC? < ch >  
Sets the readback DAC of an output < ch > to a specified number of counts.  
Queries the setting of the readback DAC of an output < ch > . The number returned is in  
DAC counts.  
Command Summary 113  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table C-2 Command Summary (continued)  
ROM?  
Queries the revision date of the power supplys firmware. See service manual.  
SROM?  
SRQ <X>  
Queries the revision date of the secondary ROM. See service manual.  
Sets the causes for generating SRQ. Setting < x > can be a 0, 1, 2, or 3 as described on  
page 79.  
SRQ?  
Queries the present setting of the reasons for issuing an SRQ (see page 79). Response is  
0, 1, 2, or that corresponds with the settings for SRQ.  
STO < x >  
Stores the present operating state of all outputs in one of eleven (0-10) registers < x > .  
Register 0 will store values of: VSET, VRSET, ISET, IRSET, OVSET, OCP, DLY, and  
MASK.  
Registers 1-10 will store values of: VSET, VRSET, ISET, IRSET, and OVSET.  
Registers 1 - 3 are contained in the EEPROM (U230) and are non-volatile. Registers 6-  
10 are located in RAM (U207) and are volatile. Whenever the supply is powered up, it  
initializes all outputs using the values stored in register 0. STO <0,1,2,3> can only be  
sent once. An error will result if an attempt is made to store more than once into each  
location. The supply must be cycled ON/OFF to re-enable.  
STS? < ch >  
Queries the present status of the specified output < ch > . The response (integer 0-255)  
represents the sum of the binary weights of the status register bits (see page 76).  
TEST?  
Causes the power supply to perform a self test of its GP-IB interface. The response is 0  
if the test passes or an integer failure code (see page 82).  
UNMASK < ch >, < x >  
Sets the bits in the mask register of the specified output channel to the setting (integer  
from 0 to 255). The mask register operates in conjunction with the status and fault  
registers (see page 78).  
UNMASK? < ch >  
VDAC < ch >, < x >  
VDAC? < ch >  
Queries the present setting of the mask register of the specified output channel (see  
page 77). The response is an integer from 0 to 255.  
Programs the voltage DAC to a specified number of counts < x > for the selected  
channel < ch >.  
Queries the LSB setting of the voltage DAC for the selected channel < ch >. The number  
returned is in DAC counts.  
VDATA< ch >, < Vlo >,  
< Vhi >  
Sends data to calibrate the voltage setting circuits of the specified output < ch > . Vlo  
and Vhi are measured values which the supply uses to calculate correction constants (see  
Appendix A).  
VHI <ch>  
Causes the voltage of the specified output < ch > to go to the high calibration point (see  
Appendix A).  
VLO < ch >  
Causes the voltage of the specified output < ch > to go to the low calibration point (see  
Appendix A).  
114 Command Summary  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table C-2 Command Summary (continued)  
Queries the measurement of the input < x > to the analog multiplexer on the specified  
output < ch > (see Appendix A).  
VMUX? < ch > , < x >  
VOUT? < ch >  
Queries the measured output voltage of the specified output < ch > (see page 70). The  
response is a real number. The front panel display can be used to monitor the measured  
output voltage (and current) of the selected output channel.  
VRDAT < ch >, < Vlo >,  
< Vhi >  
Used to calibrate the voltage readback for the specified output < ch > . This command  
must be sent after VRHI and VRLO commands.  
VRLO < ch >  
VRHI < ch >  
Finds the minimum voltage readback value for the specified output < ch > .  
Finds the maximum voltage readback value for the specified output < ch >.  
VRSET < ch > ,x  
Sets the full scale voltage range for prograrnming an output < ch > . The power supply  
will select the lowest operating range that will contain the value (x). If the present  
operating voltage of the supply is greater than the full scale range selected, the present  
output voltage will be reduced to the full scale value of the new range. If the voltage  
entered is larger than the highest possible range, an error message will result.  
VRSET < ch >  
VSET < ch >, < x >  
VSET? < ch >  
Queries the full scale voltage range setting of an output < ch >.  
Sets the voltage of the specified output < ch > (see pages 71-72).  
Queries the present voltage setting of the specified output < ch > (see page 72). The  
response is a real number.  
VSTEP< ch > , < x >  
Increases or decreases an outputs < ch > voltage in steps. If the step value < x > is  
positive, the output voltage will increase by step value; if the step value is negative, the  
output voltage will decrease by the step value. Trying to step beyond the active  
operating range will cause a number range error.  
Command Summary 115  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
D
Error Codes and Messages  
Introduction  
This appendix describes the GP-IB error codes that can be readback to the controller and the error messages that can be  
displayed on the power supplys front panel. A brief explanation of each code and message is also given. The error codes  
and/or messages fall into three categories: Power-on Self Test Messages, responses to the ERR? query, and responses to the  
TEST? query.  
Power-On Self Test Messages  
The applicable message listed in Table D-1 is displayed on the supplys front panel if the corresponding function has failed  
the power-on self test. No error code numbers can be returned for these errors. If any of these error messages appear, refer  
to the Troubleshooting section in the Service Manual.  
Error Responses  
Table D-2 describes the error codes and messages that can be generated for various programming, calibration, and hardware  
errors. The error codes can be read back over the GP-IB in response to the ERR? query. The applicable error message is  
displayed on the supplys front panel when the ERR key is pressed. Sending the ERR? or pressing the ERR key clears the  
error.  
Test Responses  
Table D-3 describes the codes that can be read back over the GP-IB in response to the TEST? query. The TEST? query  
initiates a self test of the supply.  
Table D-1. Power-On Self Test Error Messages  
Message  
TIMER FAILED  
Explanation  
The timer on the GP-IB board failed.  
8291 FAILED  
GP-IB control chip on the GP-IB board failed.  
CV DAC CH < ch >  
CC DAC CH < ch >  
OV DAC CH < ch >  
FUSE CH < ch >  
The voltage DAC on the specified output board failed.  
The current DAC on the specified output board failed.  
The overvoltage DAC on the specified output board failed.  
The return fuse on the specified output board is opened.  
The specified output board is down (no output).  
HDW ERR CH < ch >  
Table D-2. ERROR Responses  
Error Code  
Message  
Explanation  
(ERR? query)  
(ERR key)  
0
NO ERROR  
Indicates there are no errors.  
1
2
INVALID CHAR  
INVALID NUM  
You sent the supply a character it did not recognize.  
The format of your number is incorrect. Check syntax (see Chapter 5).  
Error Codes and Messages 117  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table D-2. ERROR Responses (continued)  
Explanation  
Error Code  
Message  
(ERR? query)  
(ERR key)  
3
4
5
6
7
8
9
INVALID STR  
You sent a command the supply does not understand. Resend  
recognizable command.  
SYNTAX ERROR You sent a command with improper syntax. Check syntax of your  
command (see Chapter 5).  
NUMBER RANGE An out of range number was sent. Send a new number within the legal  
range.  
NO QUERY  
The computer addressed the supply to talk, but it did not first request  
data. Send query first and then address the supply to talk.  
DISP LENGTH  
BUFFER FULL  
Quoted string in the DSP command exceeds the display length of 12  
characters.  
May occur if too many numbers are sent. Error code 4 or error code 5  
are more likely to occur for this condition.  
EEPROM ERROR The EEPROM on the GP-IB board is not responding correctly to  
programming commands. An instrument failure has occurred and  
service is required. Refer to the Troubleshooting Section in the Service  
Manual.  
10  
11  
HARDWARE ERR An output error has occurred on an unknown output. Service is required.  
Refer to the Troubleshooting Section in the Service Manual.  
HDW ERR CH 1  
Error codes 11 through 14 refer to a specific output where an output  
error has occurred. Service is required. Refer to the Output Board  
Troubleshooting Section in the Service Manual.  
12  
13  
14  
15  
HDW ERR CH 2  
HDW ERR CH 3  
HDW ERR CH 4  
Same as in error 11.  
Same as in error 11.  
Same as in error 11.  
NO MODEL NUM The supplys model number cannot be found. The GP-IB Interface board  
may be defective or the supply may require reprogramming. Refer to the  
Troubleshooting Section in the Service Manual.  
16  
17  
CAL ERROR  
An error has occurred during calibration. This may be the result of out of  
range numbers sent. If recalibration (see Appendix A) does not fix this  
there may be a hardware failure (Refer to the Troubleshooting Section in  
the Service Manual).  
UNCALIBRATED Unexplained EEPROM error; possibly as the result of incorrect  
calibration procedure. Recalibrate as described in Appendix A. If the  
problem persists, a hardware failure exists (Refer to the Troubleshooting  
Section in the Service Manual)  
118 Error Codes and Messages  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table D-2. ERROR Responses (continued)  
Explanation  
Error Code  
Message  
(ERR? query)  
(ERR key)  
18  
CAL LOCKED  
Calibration was attempted with the calibration jumper on the GP-IB  
board in the lockout position (See Section IV in the Service Manual).  
Reposition the jumper and re-calibrate if this is desired.  
22  
SKIP SLF TST  
The self test jumper on the GP-IB board is in the Skip Self Test Position  
(See Section IV in the Service Manual). Reposition the jumper and carry  
out self-test if this is desired.  
Table D3. TEST? Response  
Explanation  
Response Code  
(TEST? query)  
0
No errors were detected.  
20  
The timer on the GP-IB board failed self test. Refer to the Troubleshooting Section in the  
Service Manual.  
21  
27  
The RAM on the GP-IB Board failed self test. Refer to the Troubleshooting Section in the  
Service Manual. Note that if this failure occurred at power-on, it cannot be displayed on the  
front panel.  
The RAM on the GP-IB Board failed self test. Refer to the Troubleshooting Section in the  
Service Manual. Note that if this failure occurred at power-on, it cannot be displayed on the  
front panel.  
29  
30  
See Table 5-9 on page 84.  
See Table 5-9 on page 84.  
Error Codes and Messages 119  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
E
Manual Backdating  
Introduction  
The backdating information in this chapter applies to units that have the following serial numbers:  
Agilent Model 6625A. serials 2831A-00101 to 00663  
Agilent Model 6626A, serials 2831A-00101 to 00823  
Agilent Model 6628A, serials 2922A-00101 to 00232  
Agilent Model 6629A, serials 2920A-00101 to 00292  
Make Changes  
On page 29, replace the information in Line Voltage Conversion paragraph under steps number 2, 3, and 4 as follows:  
2. To open the line module push against the tab on the line module in the direction of the ac input socket.  
3. Remove the voltage selector card from its receptacle. This card is about one inch square and has a red notch in  
each of its four edges.  
4. To select a voltage, insert the card into its receptacle so that after the door is closed, the red notch shows through  
the window next to the voltage level you require.  
It is possible to insert the voltage selector card so that a red notch is not visible through any of the four  
windows. Do NOT operate the power supply while the selector card is in this position.  
Also on page 28, replace figure 2-3 with the following figure;  
Figure 2-3. Line Module Detail  
Manual Backdating 121  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Agilent Sales and Support Office  
For more information about Agilent Technologies test and measurement products, applications, services, and for  
a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir  
You can also contact one of the following centers and ask for a test and measurement sales representative.  
United States:  
Latin America:  
Agilent Technologies  
Test and Measurement Call Center  
P.O. Box 4026  
Englewood, CO 80155-4026  
(tel) 1 800 452 4844  
Agilent Technologies  
Latin American Region Headquarters  
5200 Blue Lagoon Drive, Suite #950  
Miami, Florida 33126  
U.S.A.  
(tel) (305) 267 4245  
(fax) (305) 267 4286  
Canada:  
Australia/New Zealand:  
Agilent Technologies Canada Inc.  
5150 Spectrum Way  
Mississauga, Ontario  
L4W 5G1  
Agilent Technologies Australia Pty Ltd  
347 Burwood Highway  
Forest Hill, Victoria 3131  
(tel) 1-800 629 485 (Australia)  
(fax) (61 3) 9272 0749  
(tel) 1 877 894 4414  
(tel) 0 800 738 378 (New Zealand)  
(fax) (64 4) 802 6881  
Europe:  
Asia Pacific:  
Agilent Technologies  
Test & Measurement European Marketing Organisation  
P.O. Box 999  
1180 AZ Amstelveen  
The Netherlands  
Agilent Technologies  
24/F, Cityplaza One, 1111 Kings Road,  
Taikoo Shing, Hong Kong  
tel: (852)-3197-7777  
fax: (852)-2506-9284  
(tel) (31 20) 547 9999  
Japan:  
Agilent Technologies Japan Ltd.  
Measurement Assistance Center  
9-1, Takakura-Cho, Hachioji-Shi,  
Tokyo 192-8510, Japan  
(tel) (81) 426 56 7832  
(fax) (81) 426 56 7840  
Technical data is subject to change.  
122 Agilent Sales and Support Office  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual Updates  
The following updates have been made to this manual since the print revision indicated on the title page.  
2/01/00  
All references to HP have been changed to Agilent.  
All references to HP-IB have been changed to GPIB.  
9/20/04  
The Declaration of Conformity has been updated.  
123  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Able Planet Fan 620DuN User Manual
ADIC Tool Storage 6 00025 01 User Manual
Agilent Technologies Clock Radio 2 User Manual
Aiphone Pet Fence HID SS User Manual
Aiphone Power Supply PS 18C User Manual
Alliance Laundry Systems Iron RI1400 25 AV User Manual
Alto Shaam Food Warmer ED2SYS 48 2S User Manual
Amana Range AEP222VAW User Manual
American Standard Air Conditioner UV05 User Manual
American Standard Hot Tub 1742 Series User Manual