Agilent Technologies Music Mixer 11970 90031 User Manual

User s Gu id e  
11970 SE RIE S  
H AR MONIC MIXE R S  
(K, A, Q, U, V, a n d W Mod els)  
Ma n u fa ct u r in g P a r t Nu m ber : 11970-90031  
Supersedes 11970-90030  
P r in t ed in USA  
Oct ob er 2003  
© Copyright 1992, 1999, 20012003 Agilent Technologies, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LIMITATION OF WAR R ANTY  
The foregoing warranty shall not apply to defects resulting from improper or inadequate  
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification  
or misuse, operation outside of the environmental specifications for the product, or  
improper site preparation or maintenance.  
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES  
SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY  
AND FITNESS FOR A PARTICULAR PURPOSE.  
EXCLUSIVE R EME DIE S  
THE REMEDIES PROVIDED HEREIN ARE BUYERS SOLE AND EXCLUSIVE  
REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT,  
INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER  
BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.  
Assist a n ce  
Product maintenance agreements and other customer assistance agreements are available  
for Agilent Technologies products.  
For any assistance, contact your nearest Agilent Technologies Sales and Service Office  
listed on page 77.  
Th is gu id e u ses th e follow in g con ven t ion s:  
Instrument Key This represents a key physically located on the instrument, or a key with a  
label that is determined by the instrument firmware.  
Screen Text This indicates text displayed on the spectrum analyzer screen.  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Con ten t s  
Using the Mixers with the E4440A, E4446A, or E4448A  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
1 Gen er a l In for m a t ion  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
General Information  
Introduction  
In t r od u ct ion  
The Agilent Models 11970K, 11970A, 11970Q, 11970U, 11970V and 11970W are  
general-purpose harmonic mixers with very flat frequency response characteristics and  
low conversion loss. Collectively, they cover the frequency range of 18 to 110 GHz. The  
11970K covers the 18 - 26. 5 GHz range; the 11970A, the 26.5 - 40 GHz range; the 11970Q,  
the 33 - 50 GHz range; the 11970U, the 40 - 60 GHz range; the 11970V, the 50 - 75 GHz  
range; and the 11970W, the 75 - 110 GHz range.  
The overall local oscillator (LO) frequency range of the 11970 Series Mixers is 3.0 to 6.1  
GHz. Each of the mixers employs a different LO harmonic, and as a result has a different  
optimum LO range within the overall LO range of the series.  
The 11970 Series Mixers use the Agilent 11975A Amplifier to raise the LO power to their  
required LO input level of 14 to 18 dBm. By taking advantage of the power leveling  
capability of the 11975A, the mixers are able to achieve maximum measurement accuracy  
(at optimum LO input level of 14.5 to 16 dBm).  
A label on the end of each mixer shows a Conversion Loss Calibration table printed  
especially for that particular mixer. An 8-1/ 2 by 11-inch calibration table shipped with the  
mixer provides a larger, easier to read, version of the same table shown on the label, plus a  
graph which shows the conversion loss and reference level offset across the mixers  
frequency range. This calibration table must be employed for absolute amplitude  
measurements. Also supplied with each mixer are five screws (four required) for attaching  
the mixer RF input flange to the waveguide.  
Mixer s Cover ed by Ma n u a l  
Ser ia l Nu m b er s  
Attached to your mixer, is a label which shows both the mixer model number and its serial  
number (two parts). The first four digits and the letter of the serial number are the serial  
number prefix; the last five digits are the suffix. The contents of this manual apply to  
mixers with the serial number prefixes listed under Serial Number Prefixeson the title  
page of this manual.  
8
Chapter1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
General Information  
Introduction  
11970 Ser ies Ha r m on ic Mixer s  
Figu r e 1-1.  
Op tion s  
Op tion 009, shown in Figure 1-2. is a Mixer Connection Kit. It includes three low-loss SMA  
cables (Part Number 5061-5458), one hex-head balldriver (Part Number 8710-1539) for  
tightening the waveguide connector screws, and one 5/16-inch open-end wrench (Part  
Number 8710-0510) for use on the SMA connectors.  
Sp ect r u m An a lyzer R et r ofit Req u ir em en t s  
The 11970 Series Mixers are fully compatible with all 8566B Spectrum Analyzers.  
856x Series portable spectrum analyzers with options 002 (tracking generator) and 327  
(no IF IN on front panel) do not have external mixing capability.  
856x Series Portable Spectrum Analyzers require option 008, or FW date code 920528,  
for signal identification functionality.  
Agilent E4407B spectrum analyzers require option AYZ (LO Out and IF IN  
connections).  
MMS analyzers using 11970 series mixers must have a front end model 70907A,  
70907B, 70909A, or 70910A, as well as a 70900A/B with firmware new enough for those  
modules:  
Model  
FW Date Code:  
(or newer)  
-
-
-
-
70907A  
70907B  
70909A  
70910A  
860203  
900314  
910802  
910802  
Chapter 1  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
General Information  
Introduction  
PSA Series E4440A, E4446A, and E4448A spectrum analyzers require Option AYZ (LO  
OUT and IF IN connectors). The E4443A and E4445A do not support external mixing.  
Figu r e 1-2. Mixer Con n ect ion Kit , Op t ion 009  
10  
Chapter1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
General Information  
Specifications  
Sp ecifica t ion s  
Specifications for the 11970 Series Mixers are listed in Table 1-1. These are the performance  
standards against which the mixers are tested (performance tests are provided in Chapter  
3 ,“Performance Tests,”). Typical or nominal operating values are listed in Table 1-2.  
NOTE  
Supplemental characteristics are included only as additional information;  
they are not specifications.  
E qu ip m en t Su p p lied  
Wa vegu id e Con n ect or Scr ew s  
Five hex-head screws are supplied with each mixer. Four are required to make the mixer-to-waveguide  
connection, one is a spare. Use ONLY the screws supplied to attach the mixer to the waveguide. Because  
of slight differences in the way the mixers couple with the waveguide, the square-flange mixers (11970K  
and 11970A) and the round-flange mixer (1970Q, 11970U, 11970V and 11970W) require different  
screws.  
If your mixer has a square waveguide flange, use four hex-head screws with Part Number  
3030-0221. If it has a round waveguide flange, use four hex-head captive screws with Part  
Number 1390-0671. The special balldriver hex screwdriver available in the Option 009  
Mixer Connector Kit simplifies installation of the waveguide connector screws.  
NOTE  
See Chapter 4 ,“Service,” for a list of replacement parts.  
En vir on m en ta l Lim ita tion s  
The 11970 Series Mixers meet or exceed the environmental requirements of MIL-T-28800C, Type III,  
Class 3, Style C. Specific environmental qualifications for the mixers are as follows:  
o
o
Temperature, Non-operating: -40 C to 75 C  
o
o
Temperature, Operating: 0 C to 55 C  
o
Relative Humidity: 95 ±5% (up to 30 C)  
Altitude, Non-operating: Less than 12,195 meters (40,000 ft.)  
Altitude, Operating: Less than 3,048 meters (10,000 ft.)  
Maximum Vibration Levels: 2 Gs at 5 to 2000 Hz  
Maximum Shock: 30 Gs  
Chapter 1  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
General Information  
Specifications  
Table 1-1 11970 Series Specifications  
NOTE  
Unless otherwise stated, all specifications apply for an IF of 321.4 MHz and for RF input amplitudes of  
less than -20 dBm.  
GENERAL  
1
Bias Requirements: None  
LO Amplitude Range: 14 to 18 dBm  
Calibration Accuracy:  
Maximum CW RF Input Level:  
20 dBm (l00mW)  
11970K/A/Q/U:  
± 2.0 dB with LO amplitude range of 14.5 to 16 dBm  
Maximum Peak Pulse Power:  
11970V/W:  
24 dBm with < 1µsec pulse (avg. power: + 20 dBm)  
± 2.2 dB with LO amplitude range of 14.5 to 16 dBm  
11970K/A/Q/U:  
Environmental:  
± 3.0 dB with LO amplitude range of 16 to 18 dBm  
Meets MIL-T-28800C, Type III, Class 3, Style C  
11970V/W: ± 3.2 dB with LO amplitude range of 16 to  
18 dBm  
IF/LO Connectors:  
SMA female  
12  
Chapter1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
General Information  
Specifications  
Table 1-1 11970 Series Specifications  
NOTE  
Unless otherwise stated, all specifications apply for an IF of 321.4 MHz and for RF input amplitudes of  
less than -20 dBm.  
MODEL 11970K  
MODEL 11970A  
RF Frequency Range:  
RF Frequency Range:  
26.5 - 40 GHz  
18 - 26.5 GHz  
LO Harmonic Number: 6  
LO Harmonic Number: 8  
LO Input Frequency Range:  
3.27 - 4.96 GHz  
LO Input Frequency Range:  
2.95 - 4.36 GHz  
Maximum Conversion Loss: 26 dB  
Maximum Conversion Loss: 24 dB  
8566B Noise Level at 1 kHz Bandwidth, and  
14.5 to 16 dBm  
LO Input Power: -102 dBm  
8566B Noise Level at 1 kHz Bandwidth, and  
14.5 to 16 dBm  
LO Input Power: -105 dBm  
Frequency Response at 14.5 to 16 dBm  
Frequency Response at 14.5 to 16 dBm  
LO Input Power: ±1.9 dB  
LO Input Power: ±1.9 dB  
Frequency Response at 14 to 18 dBm  
Frequency Response at 14 to 18 dBm  
LO Input Power: ±2.8 dB  
LO Input Power: ±2.8 dB  
MODEL 11970U  
MODEL 11970Q  
RF Frequency Range:  
RF Frequency Range:  
40 - 60 GHz  
33 - 50 GHz  
LO Harmonic Number: 10  
LO Harmonic Number: 10  
LO Input Frequency Range:  
LO Input Frequency Range:  
3.97 - 5.97 GHz  
3.27 - 4.97 GHz  
Maximum Conversion Loss: 28 dB  
Maximum Conversion Loss: 28 dB  
8566B Noise Level at 1 kHz Bandwidth, and  
14.5 to 16 dBm  
8566B Noise Level at 1 kHz Bandwidth, and  
14.5 to 16 dBm  
LO Input Power: -101 dBm  
LO Input Power: -101 dBm  
Frequency Response at 14.5 to 16 dBm  
Frequency Response at 14.5 to 16 dBm  
LO Input Power: ±1.9 dB  
LO Input Power: ±1.9 dB  
Frequency Response at 14 to 18 dBm  
Frequency Response at 14 to 18 dBm  
LO Input Power: ±2.8 dB  
LO Input Power: ±2.8 dB  
Chapter 1  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General Information  
Specifications  
Table 1-1 11970 Series Specifications  
NOTE  
Unless otherwise stated, all specifications apply for an IF of 321.4 MHz and for RF input amplitudes of  
less than -20 dBm.  
MODEL 11970W  
RF Frequency Range:  
MODEL 11970V  
RF Frequency Range:  
75 - 110 GHz  
50 - 75 GHz  
LO Harmonic Number: 14  
LO Harmonic Number: 18  
LO Input Frequency Range:  
LO Input Frequency Range:  
4.15 - 6.09 GHz  
3.55 - 5.33 GHz  
Maximum Conversion Loss: 46 dB  
Maximum Conversion Loss: 40 dB  
8566B Noise Level at 1 kHz Bandwidth, and  
14.5 to 16 dBm  
LO Input Power: -92 dBm  
8566B Noise Level at 1 kHz Bandwidth, and LO Input  
Power: - 85 dBm  
Frequency Response at 14.5 to 16 dBm  
LO Input Power: ±3.0 dB2  
Frequency Response at 14.5 to 16 dBm  
LO Input Power: ±2.1 dB  
Frequency Response at 14 to 18 dBm  
Frequency Response at 14 to 18 dBm  
LO Input Power: ±2.8 dB  
LO Input Power: ±4.0 dB  
1. The 11975A Amplifier, or a similar amplifier, can be used to provide sufficient LO power (14 to  
18 dBm) to the mixers. Leveled power capability of 16 dBm, as is available with the 11975A, is nec-  
essary to achieve the maximum amplitude accuracy with the mixers. Assuming the 1 meter cables  
supplied with the instrument are used, only the 8566B LO output requires amplification. However,  
if longer or lower quality cables are used LO amplification may be necessary.  
2. Typically ± 2.5 dB with LO supplied by 8566B Spectrum Analyzer, 11975A Amplifier set to  
16 dBm, and P/N 5061-5458 SMA cables.  
14  
Chapter1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General Information  
Specifications  
Table 1-2 11970 Supplemental Characteristics  
NOTE  
Supplemental characteristics are included only as additional information; they are not specifications.  
3 dB IF Bandwidth: DC to 1.3 GHz  
Odd Order Mixing Product Suppression:  
11970K/A/Q/U: >20 dB  
11970V/W: >15 dB  
Spectrum Analyzer Absolute Amplitude  
Accuracy (using calibration data with a 14.5 to  
16 dBm LO):  
Gain Compression Level (<1 dB):  
11970K: -3 dBm  
11970A: -5 dBm  
11970K, 18 - 26.5 GHz: ±3.2 dB  
11970A, 26.5 - 40 GHz: ±3.2 dB  
11970Q, 33 - 50 GHz: ±3.2 dB  
11970Q: -7 dBm  
11970U: -7 dBm  
11970V: -3 dBm  
11970W: -1 dBm  
11970U, 40 - 60 GHz: ±3.2 dB  
11970V, 50 - 75 GHz: ±3.4 dB  
11970W: 75 - 110 GHz: ±3.4 dB  
RF Input SWR:  
11970K/A/Q/U: <2.2:1  
11970V/W: <2.6:1  
5061-5458 Cable Insertion Loss:  
.8 dB at 2 GHz  
1.1 dB at 6 GHz  
PHYSICAL CHARACTERISTICS  
1
Z
Model  
Weight  
X
Y
Flange  
0.17 kg  
0.36 lb  
51 mm  
2.0 in  
90 mm  
3.5 in  
11970K  
UG-595/U  
WR-42  
36 mm  
1.4 in  
36 mm  
1.4 in  
51 mm  
2.0 in  
71 mm  
2.8 in  
11970A  
11970Q  
11970U  
11970V  
11970W  
UG-599/U  
WR-28  
0.14 kg  
0.32 lb  
UG-383/U  
WR-22  
0.14 kg  
0.32 lb  
51 mm  
2.0 in  
36 mm  
1.4 in  
76 mm  
3.0 in  
UG-383/U-M  
WR-19  
76 mm  
3.0 in  
0.14 kg  
0.32 lb  
36 mm  
1.4 in  
51 mm  
2.0 in  
UG-385/U  
WR-15  
0.14 kg  
0.32 lb  
36 mm  
1.4 in  
51 mm  
2.0 in  
76 mm  
3.0 in  
UG-387/U-M  
WR-10  
0.14 kg  
0.32 lb  
51 mm  
2.0 in  
76 mm  
3.0 in  
36 mm  
1.4 in  
1. Waveguide attachment screws enter blind holes in the flanges of the mixers.  
Chapter 1  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
General Information  
Specifications  
16  
Chapter1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2 Op er a tion  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation  
Introduction  
In tr od u ction  
This section provides information on how to make effective use of the 11970 mixers.  
Op er a t in g P r eca u tion s  
Refer to the sections below for specific parameters to follow prior to mixer operation.  
WARNING  
Do n ot exceed t h e m a xim u m r a t in gs list ed below or p er m a n en t  
d a m a ge t o t h e m ixer w ill r esu lt .  
RF In p u t P ow er  
Use the following parameters:  
CW: No greater than 20 dBm  
Pulse: No greater than 24 dBm at < 1 mSec  
Average: No greater than 20 dBm  
LO In p u t Pow er  
Make sure the LO input power is no greater than 20 dBm.  
E lect r ost a t ic Disch a r ge  
When installing the mixer, you must always connect the SMA cables to the spectrum  
analyzer and LO amplifier BEFORE connecting them to the mixer. This will minimize the  
danger of an electrostatic discharge damaging the mixer diodes.  
11975A ALC Sw it ch  
BEFORE using the 11975A Amplifier to increase the LO input power, set the amplifier  
ALC switch to the ON position. When this switch is in the OFF position the LO power can  
be greater than 20 dBm. This level of LO power can destroy the mixer diodes. The ALC  
switch is on the amplifier rear panel.  
Wa vegu id e P r ot ect ive Foa m  
Do not remove, displace, or damage the white, nonconductive foam installed in the open  
end of the waveguide. Since the mixer is amplitude calibrated with this foam in place,  
tampering with it affects the calibration.  
18  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operation  
Getting Started  
Gettin g Sta r ted  
The 11970 series of millimeter wave mixers have no bias or back-short adjustments. 11970  
Mixers require an LO power of 14 to 18 dBm at the LO input. If the spectrum analyzer  
used with the 11970 does not have sufficient LO power, an 11975A Amplifier or an  
equivalent is required to increase this power.  
CAUTION  
Before connecting the 11975A Amplifier, set its rear panel ALC switch to ON.  
Failure to do this can damage the mixer.  
Mixer Con n ection s  
With the three SMA cables (Part Number 5061-5458, each) provided in the Option 009  
Mixer Connection Kit, connect the 11970 Mixer, the spectrum analyzer, and the 11975A  
Amplifier. An example connection for an 8566B Spectrum Analyzer is shown in Figure 2-1.  
NOTE  
The example below shows an 8566B analyzer, other spectrum analyzers do  
not require the amplifier and third SMA cable.  
Figu r e 2-1. 11970 Mixer Con n ect ion s  
Leave the waveguide flange cap on whenever the mixer is not connected to a device under  
test. This protects the flange mating surface from scratches, which can degrade the mixers  
performance. Use an appropriate waveguide attenuator if the output power of the unit  
under test exceeds the RF Input Power indicated in the specifications.  
Chapter 2  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operation  
Getting Started  
If you are using an 11970Q, 11970U, 11970V or 11970W Mixer, and the shoulder of its waveguide  
flange is not properly aligned with the flange of the device under test, amplitude measurement errors can  
result. To ensure proper alignment, tighten each of the four flange screws in turn by small amounts,  
moving clockwise around the flange.  
NOTE  
11970K and 11970A Mixers require flange screws that are different from  
those used with the 11970Q, 11970U, 11970V and 11970W; 11970K and  
11970A use Part Number 3030-0221 (#4-40 hex head screw). 11970Q,  
11970U, 11970V and 11970W use Part Number 1390-0671 (#4-40 hex head  
captive screw).  
20  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Using a Conversion-Loss Data Disk with the ESA or PSA Series Analyzers  
Usin g a Con ver sion -Loss Da ta Disk w ith th e ESA or P SA  
Ser ies An a lyzer s  
The conversion-loss data for your mixer can be quickly loaded in the memory of an ESA or  
PSA series analyzer from a floppy disk. The following process works with the floppy disk  
provided by the factory with your new mixer, or with one that you have stored the  
correction data. This process eliminates the time consuming and sometimes error-prone  
process of manually entering the data from a table.  
Loa d in g Da ta  
1. Insert the conversion-loss data disk (from the factory or your own) into the floppy drive  
of the analyzer.  
2. Press File, then Load.  
3. Press Type, More, then select Corrections.  
4. Press Directory Select, highlight drive -A-, then press Directory Select again.  
5. Use the down arrow to select the file. The factory supplied disk has a file with the OTH  
extension which places the data in the corrections set otherlocation.  
6. Press Load Now to load the conversion loss values into the correction table of the  
analyzer.  
View in g Da ta  
1. Press Amplitude, More, Corrections, Other (ON) and Edit to view the correction values.  
2. Use the front-panel knob or up/down arrows to scroll through the list of correction  
values.  
Chapter 2  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Operation  
Using the Mixers with the E4407B Spectrum Analyzer (Option AYZ)  
Usin g th e Mixer s w ith th e E4407B Sp ectr u m An a lyzer  
(Op tion AYZ)  
The Agilent Technologies E4407B spectrum analyzer contains an extensive menu of  
functions that help with millimeter measurements. The following examples explain how to  
connect external mixers to the spectrum analyzer, how to choose the band of interest, how  
to store and activate conversion-loss factors, and how to use the signal-identification  
functions.  
Set u p th e equ ip m en t  
1. Connect the signal source and harmonic mixer to the analyzer, as shown in Figure 2-2.  
Figu r e 2-2 Ha r m on ic Mixer Set u p  
NOTE  
Agilent 5061-5458 SMA type cables should be used to connect the mixer IF  
and LO ports to the analyzer. Do not over-tighten the cables. The maximum  
torque should not exceed 112 N-cm (10 in-lb.)  
22  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operation  
Using the Mixers with the E4407B Spectrum Analyzer (Option AYZ)  
2. On the analyzer, press Preset, Factory Preset, if present.  
3. Select external mixing by pressing Input/Output, Input Mixer, Input Mixer (Ext).  
4. The analyzer frequency band will be set to 26.5 - 40 GHz (A). To choose a different band,  
press Ext Mix Band , then press the desired band frequency range/letter key. For this  
example, we will use band A, which ranges from 36.5 GHz to 40 GHz.  
Am p lit u d e Ca lib r a t ion  
The conversion loss versus frequency data can be entered on your analyzer from one of  
three sources.  
From a Conversion Loss Data Disk, supplied with your mixer. See Using a  
Data from the calibration sheet supplied with your mixer.  
From conversion-loss data located on the mixer body label.  
Ma n u a lly En t er in g Con ver sion -Loss Da t a  
1.The analyzer frequency band will be set to 26.5 – 40 GHz (A). To choose a different band,  
press Ext Mix Band and then press the desired band frequency range/letter key. For this  
example, we will use band A, which ranges from 26.5 GHz to 40 GHz.  
NOTE  
To correct for the conversion-loss of the harmonic mixer in use, the analyzer  
amplitude correction feature is used.  
2. Press AMPLITUDE Y Scale, More , Corrections. Select a correction set for use with  
external mixing. The recommended set to use is Other although any available set could  
be used.  
3. Press Edit to enter the appropriate conversion loss data for the mixer in use. These  
values are listed on the mixer, or a calibration sheet that is supplied with the mixer.  
NOTE  
More correction points entered across the band in use will improve frequency  
response accuracy. Up to 200 points may be defined for each set.  
4. Once the desired correction points are entered, press Return, Correction (On) to activate  
correction set Other. This will also turn corrections on resulting in a calibrated display.  
It is recommended that the correction set entered be saved on the internal memory or  
the floppy drive for future reference. See the ESA Users and Programmers Reference  
guide for information on saving correction values.  
Chapter 2  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Using the Mixers with the E4407B Spectrum Analyzer (Option AYZ)  
Sign a l Id en tifica t ion  
The IF output of a harmonic mixer will contain a signal at the intermediate frequency of the  
analyzer whenever the harmonic frequency of the LO and the frequency of the RF differ by the  
intermediate frequency.  
As a result, within a single harmonic band, a single input signal can produce multiple responses on  
the analyzer display, only one of which is valid (see Figure 2-3.). These responses come in pairs,  
where members of the valid response pair are separated by 642.8 MHz and either the right-most  
(for negative harmonics) or left-most (for positive harmonics) member of the pair is the correct  
response.  
Figu r e 2-3.  
Identification of valid responses is achieved by simply turning on the signal-identification feature.  
(instrument preset selects the Image Suppress signal identification mode.) Press Input/Output, Input  
Mixer, Signal Ident (On) and note that now only the valid response (35 GHz) remains.  
Press Peak Search to place a marker on the remaining response. The signal-identification routine  
can introduce slight amplitude errors which is indicated by the message Signal Ident On,  
Amptd Uncal. Refer to Figure 2-4..  
After identifying a signal of interest, press Signal Ident (Off) before making final amplitude  
measurements. Note that Image Suppress should only be used to identify the fundamental signal  
and not for accurate amplitude measurements.  
24  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation  
Using the Mixers with the E4407B Spectrum Analyzer (Option AYZ)  
Figu r e 2-4.  
Chapter 2  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation  
Using the Mixers with the E4440A, E4446A, or E4448A PSA Series Spectrum Analyzer (Option  
AYZ)  
Usin g th e Mixer s w ith th e E4440A, E4446A, or E4448A  
P SA Ser ies Sp ectr u m An a lyzer (Op tion AYZ)  
requirements for the PSA Series spectrum analyzers. The Agilent Technologies PSA Series  
spectrum analyzers contain an extensive menu of functions that help with millimeter  
measurements. The following examples explain how to connect external mixers to the  
spectrum analyzer, how to choose the band of interest, how to store and activate  
conversion-loss factors, and how to use the signal-identification functions.  
Set u p th e equ ip m en t  
1. Connect the signal source and harmonic mixer to the analyzer, as shown in Figure 2-2.  
Figu r e 2-5 Ha r m on ic Mixer Set u p  
Spectrum Analyzer  
SMA Cable  
SMA Cable  
IF INPUT  
1st LO OUTPUT  
IF  
LO  
RF Input  
Agilent 11970 SERIES  
HARMONIC MIXER  
SIGNAL  
SOURCE  
unpremix5  
NOTE  
Agilent 5061-5458 SMA type cables should be used to connect the mixer IF  
and LO ports to the analyzer. Do not over-tighten the cables. The maximum  
torque should not exceed 112 N-cm (10 in-lb.)  
26  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operation  
Using the Mixers with the E4440A, E4446A, or E4448A PSA Series Spectrum Analyzer (Option  
AYZ)  
2. On the analyzer, press Preset, Factory Preset, if present.  
3. Select external mixing by pressing Input/Output, Input Mixer, Input Mixer (Ext).  
4. The analyzer frequency band will be set to 26.5 - 40 GHz (A). To choose a different band,  
press Ext Mix Band , then press the desired band frequency range/letter key. For this  
example, we will use band A, which ranges from 36.5 GHz to 40 GHz.  
Am p lit u d e Ca lib r a t ion  
The conversion loss versus frequency data can be entered on your analyzer from one of  
three sources.  
From a Conversion Loss Data Disk, supplied with your mixer. See Using a  
Data from the calibration sheet supplied with your mixer.  
From conversion-loss data located on the mixer body label.  
Ma n u a lly En t er in g Con ver sion -Loss Da t a  
1.The analyzer frequency band will be set to 26.5 40 GHz (A). To choose a different band,  
press Ext Mix Band and then press the desired band frequency range/letter key. For this  
example, we will use band A, which ranges from 26.5 GHz to 40 GHz.  
NOTE  
To correct for the conversion-loss of the harmonic mixer in use, the analyzer  
amplitude correction feature is used.  
2. Press AMPLITUDE Y Scale, More , Corrections. Select a correction set for use with  
external mixing. The recommended set to use is Other although any available set could  
be used.  
3. Press Edit to enter the appropriate conversion loss data for the mixer in use. These  
values are listed on the mixer, or a calibration sheet that is supplied with the mixer.  
NOTE  
More correction points entered across the band in use will improve frequency  
response accuracy. Up to 200 points may be defined for each set.  
4. Once the desired correction points are entered, press Return, Correction (On) to activate  
correction set Other. This will also turn corrections on resulting in a calibrated display.  
It is recommended that the correction set entered be saved on the internal memory or  
the floppy drive for future reference. See the ESA Users and Programmers Reference  
guide for information on saving correction values.  
Chapter 2  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Using the Mixers with the E4440A, E4446A, or E4448A PSA Series Spectrum Analyzer (Option  
AYZ)  
Sign a l Id en tifica t ion  
The IF output of a harmonic mixer will contain a signal at the intermediate frequency of the  
analyzer whenever the harmonic frequency of the LO and the frequency of the RF differ by the  
intermediate frequency.  
As a result, within a single harmonic band, a single input signal can produce multiple responses on  
the analyzer display, only one of which is valid (see Figure 2-6.). These responses come in pairs,  
where members of the valid response pair are separated by 642.8 MHz and either the right-most  
(for negative harmonics) or left-most (for positive harmonics) member of the pair is the correct  
response.  
Figu r e 2-6.  
Identification of valid responses is achieved by simply turning on the signal-identification feature.  
(instrument preset selects the Image Suppress signal identification mode.) Press Input/Output, Input  
Mixer, Signal Ident (On) and note that now only the valid response (35 GHz) remains.  
Press Peak Search to place a marker on the remaining response. Refer to Figure 2-7.  
After identifying a signal of interest, press Signal Ident (Off) before making final amplitude  
measurements. Note that Image Suppress should only be used to identify the fundamental signal  
and not for accurate amplitude measurements.  
28  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation  
Using the Mixers with the E4440A, E4446A, or E4448A PSA Series Spectrum Analyzer (Option  
AYZ)  
Figu r e 2-7.  
Chapter 2  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation  
Using the Mixers with the 856X Series Spectrum Analyzers  
Usin g th e Mixer s w ith th e 856X Ser ies Sp ectr u m An a lyzer s  
External millimeter mixers can be used to extend the frequency coverage of the 8560  
E-Series and EC-Series spectrum analyzers. (The 8560E/EC Option 002 and Option 327 do  
not have external mixing capability.)  
The 8560 E-Series and EC-Series spectrum analyzers contain an extensive menu of  
functions that help with millimeter measurements. This example explains how to connect  
external mixers to the spectrum analyzer, how to choose the band of interest, how to store  
conversion-loss factors, and how to use the optional automatic signal-identification  
functions.  
Set u p th e equ ip m en t  
1. Figure 2-8 illustrates how to connect an external harmonic mixer to the spectrum  
analyzer.  
Figu r e 2-8 Ext er n a l Mixer Set u p  
NOTE  
Good-quality shielded SMA-type cables should be used to connect the mixer to  
the spectrum analyzer to ensure that no signal attenuation occurs. Agilent  
5061-5458 SMA-type cables may be used. Do not over-tighten the cables; the  
maximum torque should not exceed 112 N-cm (10 in-lb.).  
30  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operation  
Using the Mixers with the 856X Series Spectrum Analyzers  
Select th e Fr eq u en cy Ba n d  
2. Specify unpreselected external mixing by pressing CONFIG, then EXT MXR PRE UNPR  
until UNPRis selected.  
3. To select a frequency above 18 GHz:  
a. Press AUX CTRL, EXTERNAL MIXER to set the analyzer to external mixer mode.  
b. Enter the desired frequency directly using the center-frequency function.  
c. Notice in Table 2-1 that some frequencies overlap and fall into two bands. To be sure  
that the desired band is selected, refer to Table 2-1 and select the desired frequency  
band, then use the full band function to enter this band.  
d. In the external mixer menu, press FULL BAND, then press the step up key until the  
letter preceding BANDin the active function area corresponds to the desired  
frequency band.  
Table 2-1 External Mixer Frequency Ranges  
F r eq u en cy F r eq u en cy  
Mixin g  
H a r m on ic  
Con ver sion  
Loss  
Ba n d  
R a n ge (GH z)  
K
A
Q
U
V
E
18.0 to 26.5  
26.5 to 40.0  
33.0 to 50.0  
40.0 to 60.0  
50.0 to 75.0  
60.0 to 90.0  
75.0 to 110.0  
90.0 to 140.0  
110.0 to 170.0  
140.0 to 220.0  
170.0 to 260.0  
220.0 to 325.0  
6−  
30 dB  
30 dB  
30 dB  
30 dB  
30 dB  
30 dB  
30 dB  
30 dB  
30 dB  
30 dB  
30 dB  
30 dB  
8−  
10−  
10−  
14−  
16−  
18−  
24−  
30−  
36−  
44−  
54−  
W
F
D
G
Y
J
Chapter 2  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operation  
Using the Mixers with the 856X Series Spectrum Analyzers  
In this example, well look at U-band, which ranges from 40 GHz to 60 GHz, as shown in  
Figure 2-9. The LOCK HARMONIC function locksthe spectrum analyzer in that band,  
ensuring that the spectrum analyzer sweeps only the chosen band. LOCK ON OFF is  
automatically set to ON when FULL BAND is pressed.  
Figu r e 2-9 Select t h e ba n d of in t er est .  
Sa ve th e a ver a ge con ver sion -loss va lu e  
4. Default conversion-loss values that are stored in the analyzer for each frequency band  
are listed in Table 2-1. These values approximate the values for the Agilent 11970 series  
mixers. Other conversion-loss values may be entered into the spectrum analyzer in two  
ways. The first method lets you to save the average conversion-loss value for the entire  
band using AVERAGE CNV LOSS. To activate this function:  
a. Press AUX CTRL, EXTERNAL MIXER, AMPTD CORRECT, AVERAGE CNV LOSS.  
b. Enter the appropriate average conversion-loss value. On Agilent 11970 Series  
harmonic mixers, these values are printed on the mixer. The U-band mixer used for  
this example had an average conversion loss of 23.5 dB, as shown in Figure 2-10.  
32  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operation  
Using the Mixers with the 856X Series Spectrum Analyzers  
Figu r e 2-10 St or e a n d cor r ect for con ver sion loss.  
The second method for storing conversion-loss information lets you save individual  
conversion-loss data points at specific intervals across the harmonic band, using CNV  
LOSS VS FREQ.  
To view or enter a conversion-loss data point:  
a. Press CNV LOSS VS FREQ.  
b. Enter the conversion-loss data at the frequency shown.  
c. Use the step up key to step through the band, entering the conversion loss at each  
step.  
Sign a l Id en tifica tion  
5. The IF output of a harmonic mixer contains many mixer products (frequencies of LO ±  
source, 2LO ± source, 3LO ± sourcenLO ± source). As a result, within a single  
harmonic band, a single input signal can produce many responses, only one of which is  
valid.  
These responses come in pairs, where the members of the valid pair are spaced 621.4  
MHz apart (see Figure 2-11) and the right-most member for the pair is the correct  
response (for this analyzer, the left member of a pair is not valid).  
Chapter 2  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operation  
Using the Mixers with the 856X Series Spectrum Analyzers  
Figu r e 2-11 Sign a l Resp on ses P r od u ced b y a 50 GHz Sign a l in U Ba n d  
Id en tify sign a ls w ith th e fr equ en cy-sh ift m eth od  
6. Signal-identification routines that identify the signal and images are available on  
instruments with firmware revisions 920528, or with Option 008. The frequency-shift  
method of identifying valid signals uses the spectrum-analyzer function SIG ID ON OFF.  
When using this function, smaller spans will yield more accurate measurements.  
a. Span down to 50 MHz.  
b. Press AUX CTRL, EXTERNAL MIXER, SIGNAL IDENT.  
c. Press SIG ID ON OFF until ON is selected.  
Any signal not produced by the currently selected harmonic will be shifted horizontally  
on alternate sweeps (see Figure 2-12). Therefore, when viewing the display, on alternate  
sweeps the signal will appear, then not appear, at the frequency of interest. The correct  
signal produced by the selected harmonic will be shifted in a vertical direction only, as  
shown in Figure 2-13.  
34  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operation  
Using the Mixers with the 856X Series Spectrum Analyzers  
Figu r e 2-12 Resp on se for In va lid Sign a ls  
Figu r e 2-13 Resp on se for Va lid Sign a ls  
Chapter 2  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operation  
Using the Mixers with the 856X Series Spectrum Analyzers  
Id en t ify sign a ls in w id e fr eq u en cy sp a n s  
7. SIG ID AT MKR identifies signals in wide frequency spans, using harmonic search. SIG ID  
AT MKR automatically determines the proper frequency of a signal and displays its  
value on the spectrum analyzer.  
Activating SIG ID AT MKR on an image of the signal will yield a reading in the active  
block, as shown in Figure 2-14. The true signal frequency is given with an identifier  
that the marker resides on an image. If the marker is placed on a true signal, and  
SIG ID AT MKR is activated, the signal frequency will appear without the IMAGE  
notation, as shown in Figure 2-15.  
8. To activate SIG ID AT MKR:  
a. Place a marker on a signal.  
b. Press AUX CTRL, EXTERNAL MIXER, SIGNAL IDENT, SIG ID AT MKR.  
9. To exit the external mixer mode, press AUX CTRL, then INTERNAL MIXER.  
Figu r e 2-14 SIG ID AT MKR P er for m ed on a n Im a ge Sign a l  
36  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operation  
Using the Mixers with the 856X Series Spectrum Analyzers  
Figu r e 2-15 SIG ID AT MKR P er for m ed on a Tr u e Sign a l  
Chapter 2  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation  
Using the Mixers with MMS Analyzers  
Usin g th e Mixer s w ith MMS An a lyzer s  
P r elim in a r y Op er a tion  
NOTE  
This section provides information for the Agilent 70907A (or B), however the  
operation of the Agilent 70909A and Agilent 70910A is similar. Minor  
differences may be noted, but the necessary deviations from these exact  
instructions will be obvious.  
Eq u ip m en t Con n ect ion  
Connect the equipment as shown in Figure 2-16 Use the following checklist to verify the  
connections:  
Agilent 11970 LO IN to Agilent 70907B LO OUTPUT (SMA cable, 5061-5458)  
Agilent 11970 IF OUT to Agilent 70907B IF INPUT (SMA cable, 5061-5458)  
Figu r e 2-16 Agilent 11970 Connections to the Agilent 71000 Series Spectrum Analyzer  
38  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operation  
Using the Mixers with MMS Analyzers  
Op er a t ion  
Ba n d Select ion  
Use the following key sequence to enter the external mixing mode and to select the desired  
band of operation:  
MENU  
Select Input - choose external mixer input, for example: IN 2 EM 70910A”  
State  
ext mixer  
fulband KAQUVE  
Press the softkey for the desired band.  
NOTE  
The above band selection key sequence must be used to enter the external  
mixing band of operation. Do not enter in the center frequency of the band  
directly.  
Am p lit u d e Ca libr a t ion for a Sin gle F r equ en cy  
Enter the conversion loss at the desired frequency from either the calibration label on the  
end of the Agilent 11970 or the supplied calibration sheet by using the following key  
sequence:  
State  
ext mixer  
CONV LOSS  
Enter the conversion loss value of the desired frequency, then terminate the entry with dB.  
NOTE  
The above procedure is a single-point conversion-loss correction. For an  
alternative procedure, available by remote programming only, refer to  
Chapter 2  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation  
Using the Mixers with MMS Analyzers  
Gen er a l Descr ip t ion s of Agilen t 71000 Ser ies Sp ect r u m An a lyzer  
Ext er n a l-Mixin g F u n ct ion s  
ext mixer  
Allows access to the following softkey functions that control the  
measurement range when an external mixer extends the  
spectrum analyzer frequency range. The ext mixer softkey can be  
found under the State key.  
fulband KAQUVE  
fulband WFDGYJ  
CONV LOSS  
SIGNAL IDENT  
fulband KAQUVE  
and  
fulband WFDGYJ  
Specify the frequency range for measurements made with  
external mixers. Twelve frequency bands are available. Table 2-2  
lists the start and stop frequencies of each band and also shows  
which local-oscillator harmonic (N) is used for the mixing process.  
Once a band is selected, the frequency range is locked and the  
spectrum analyzer tunes with the local-oscillator harmonic for  
that band only.  
Table 2-2 External Mixing Bands and Associated Mixing Harmonics  
Ba n d  
Ha r m on ic Nu m b er (N)  
Fr eq u en cy R a n ge (GHz)  
K
A
Q
U
V
E
6−  
8+  
18.0 to 26.5  
26.5 to 40.0  
33.0 to 50.0  
40.0 to 60.0  
50.0 to 75.0  
60.0 to 90.0  
75.0 to 110.0  
90.0 to 140.0  
110.0 to 170.0  
140.0 to 220.0  
170.0 to 260.0  
220.0 to 330.0  
10+  
10+  
14+  
16+  
18+  
24+  
30+  
36+  
42+  
50+  
W
F
D
G
Y
J
40  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation  
Using the Mixers with MMS Analyzers  
CONV LOSS  
(conversion loss)  
Offsets the reference level to compensate for amplitude losses at  
the active input port. If necessary, use select input to activate the  
desired input port before specifying its conversion-loss offset.  
To clear the offset, enter a conversion loss of zero.  
After the instrument preset, a default value of 30 dB is activated  
automatically for the input port of the external-mixer-interface  
module.  
Preset Input  
Sets which active input port will be selected when the INSTR  
PRESET is pressed. The preset input softkey can be accessed by  
pressing State, then MORE.  
INSTR PRESET  
Resets to the preset input and to the lowest full frequency range  
available. (For an external mixer, this is band A.) This is a front  
panel key.  
SIGNAL IDENT  
(signal identify)  
Press Frequency, More, sig id options.  
Executes a routine to identify real, in-band signals using either  
the image (default) or shift methods.  
Several options are available under sig id options:  
- SIG ID CF  
Places the marker on the identified real signal and  
centers the signal on the display.  
- SIG ID MAN OFF  
- IMAGE/SHIFT  
Selects manual or automatic (default) mode.  
Selects the routine to use for identification.  
- SIG ID AMPTD ∆  
Specifies the amplitude difference between the  
marked signal and signals tested as images or  
harmonics (default is 10 dB; applies only to IMAGE  
mode.)  
- IMAGE N START  
Determines the frequency range for possible image or  
harmonic responses by selecting the lowest and  
highest local oscillator harmonic used for mixing,  
according to the tuning equation below:  
F = N x F  
± IF  
LO  
s
Default is 1.  
- IMAGE N STOP See IMAGE N STARTdefinition above.  
Default is 40.  
Chapter 2  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Using the Mixers with MMS Analyzers  
NOTE  
See the table below for available frequency ranges and related harmonic numbers. Use  
the IMAGE N START and IMAGE N STOP softkeys for the image identification only.  
Table 2-3 Frequency Ranges and Corresponding Harmonic Numbers  
Ba n d /R a n ge  
Ha r m on ic Nu m b er a n d Sign of  
IF (N)  
2.9 - 6.2  
6.0 - 12.7  
12.5 - 19.9  
1  
2  
+3  
+4  
In ter n a l Mixin g - Agilen t  
71210C  
19.7 - 22.0/26.5  
2.9 - 6.2  
6.0 - 12.8  
12.6 - 22.0  
1  
2  
4  
In ter n a l Mixin g - Agilen t  
71200C  
E xter n a l Mixin g - Agilen t  
70907A  
K 18.0 - 26.5  
A 26.5 - 40.0  
+6  
+8  
7090  
Q
U
33.0 - 50.0  
40.0 - 60.0  
+10  
+10  
+14  
+16  
+18  
+24  
+30  
+36  
+42  
+50  
7B  
709  
709  
V 50.0 - 75.0  
10A  
09A  
E
W
F
D
G
60.0 - 90.0  
75.0 - 110.0  
90.0 - 140.0  
110.0 - 170.0  
140.0 - 220.0  
Y 170.0 - 260.0  
220.0 - 330.0  
J
42  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Using the Mixers with MMS Analyzers  
Con ver sion Loss Ver su s F r equ en cy Cor r ection  
The Agilent 71000 Series Spectrum Analyzers with the Agilent 70907A (or B) External  
Mixer Interface Module installed has the amplitude-correction function (AMPCOR)  
available by remote programming.  
Use AMPCOR to compensate for the Agilent 11970 Series frequency-dependent  
conversion-loss variations. Up to 200 pairs of frequency-amplitude correction points can be  
entered, depending on the amount of available internal memory. The frequency values  
entered must be in increasing order, or an error condition results. Whenever AMPCOR is  
on, the correction values are added to all measurement results.  
The values of the correction points are applied across the active measurement range.  
Between points, the correction value are interpolated. When the measuring at frequencies  
outside of the first and last correction pairs from memory.  
When AMPCOR is executed, the frequency pairs are stored in internal memory. The  
remote commands DISPOSE ALL, FORMAT, ERASE, and DISPOSE AMP CORerase the  
amplitude-correction pairs from memory.  
NOTE  
DISPOSE ALL, FORMAT, and ERASEerases much more than the  
amplitude-correction pairs from memory. Do not use these commands without  
referring to the programmers manual for the Agilent 70000 and Agilent  
70900B to determine their effect on instrument programming.  
Because AMPCOR adjusts IF gain and attenuation, AMPCOR should be used with care  
when measuring signals near the compression level (also at a frequency that is corrected  
by the AMPCOR command).  
To use AMPCOR for the Agilent 11970 conversion loss correction, first set the single-point  
conversion loss function to zero with the following command:  
OUTPUT 718; “CNVLOSS 0DB;”  
Then build the AMPCOR table with frequency conversion loss pairs. The following  
example program line shows three pairs of frequency conversion loss values for a typical  
Agilent 11970V.  
OUTPUT 718; “AMPCOR 50GHZ, 49.6DB, 51GHZ, 49.1DB, 52GHZ,48.6DB;”  
The conversion loss corrections are activated by the AMPCOR ON command; use  
AMPCOR OFF to deactivate the function (be sure to manually enter a single-point  
conversion-loss value if AMPCOR has been used).  
Chapter 2  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operation  
Using the Mixers with MMS Analyzers  
To print out an existing AMPCOR table, use the program listed below:  
10 dim A$ (1:20) [30]  
20 OUTPUT 718; CONVLOSS 0DB;”  
30 OUTPUT 718; AMPCOR 50 GHZ, 46.1DB, 52GHZ, 46.7DB, 53GHZ,47.2DB;”  
40 !  
50 OUTPUT 718; AMPCOR?;”  
60 ENTER 718 USING %,k;A$ (*)  
70 !  
80 FOR I=1 TO 10  
90 PRINT Point ;I,A$ (I)  
100 NEXT I  
110 END  
Lin e 30:  
Lin e 60:  
Generates an AMPCOR table.  
% specifies that an EOI terminates the entire statement. Kspecifies that  
an  
L terminates the individual string entry.  
F
Returned values are in the format where the frequency in Hz is listed first, followed by the  
conversion loss in dB. See the following example:  
Point 1 50.00000000E+9,46.1  
Point 2 52.00000000E+9,46.7  
Point 3 53.00000000E+9,47.2  
Point 4  
Point 5  
Point 6  
Point 7  
Point 8  
Point 9  
Point 10  
Point 11  
NOTE  
For further information, consult the Agilent 70900B  
Local-Oscillator-Controlled Modules Programming Manual, part number  
70900-90284.  
44  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Operation  
Using the Mixers with the 8566B Spectrum Analyzer  
Usin g th e Mixer s w ith th e 8566B Sp ectr u m An a lyzer  
Set u p th e equ ip m en t  
1. Connect the external harmonic mixer to the spectrum analyzer, as shown in Figure 2-17.  
NOTE  
Good-quality shielded SMA-type cables should be used to connect the mixer to  
the spectrum analyzer to ensure that no signal attenuation occurs. Agilent  
5061-5458 SMA-type cables may be used. Do not over-tighten the cables; the  
maximum torque should not exceed 112 N-cm (10 in-lb.).  
Figu r e 2-17 External Mixer Connection  
CAUTION  
Before connecting the Agilent 11975A Amplifier, set the ALC switch to ON.  
Failure to do so can damage the mixer.  
2. Power up the spectrum analyzer and then the amplifier.  
3. Adjust the amplifier power to 16 dBm.  
4. To access the spectrum analyzers millimeter bands, press SHIFT and then the up arrow  
until the desired band is displayed on the screen, as shown in Figure 2-18. Note that  
this is important because toggling through the bands, causes the band to automatically  
harmonic lock as evidenced by the on-screen annotation of Harmonic 6changing to  
Chapter 2  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operation  
Using the Mixers with the 8566B Spectrum Analyzer  
Harmonic 6L, where Lstands for locked, or you could press Shift+z to manually  
harmonic lock.  
Figu r e 2-18 TYPICAL MILLIMETER SIGNAL DISPLAY  
A menu of millimeter measurement functions is accessed by pressing SHIFT, 1, MHz. Even  
though the spectrum analyzer is in an external mixer band, the 8566B spectrum analyzer  
retains full operation capability (except for use of the RF input attenuator and preselector).  
Am p lit u d e Ca lib r a tion  
P r oced u r e  
Each 11970 series harmonic mixer is characterized at the factory and comes with a  
calibration chart (See Figure 2-19). The horizontal axis of the chart gives the frequency  
range of the mixer and the two vertical axis scales are calibrated in conversion loss and in  
reference level offset.  
To correct the amplitude calibration of the spectrum analyzer to account for the conversion  
loss of the harmonic mixer, press SHIFT, 1, MHz, which accesses the millimeter menu. Then  
press SHIFT, 5, MHz to access conversion loss entry and key in the appropriate value on the  
spectrum analyzer as shown on the mixer calibration chart.  
For measurements that span the entire band, choose an average value of conversion loss  
from the conversion loss chart. For those spectrum analyzers with date codes earlier than  
13.8.86, the conversion loss is changed using the reference level function. Press SHIFT, then  
REFERENCE LEVEL, then key in the reference level offset shown on the right side of the  
mixer calibration chart.  
46  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Operation  
Using the Mixers with the 8566B Spectrum Analyzer  
Sign a l Id en tifica tion  
Figure 2-18 shows a typical full-band display of a single input signal. Direct interpretation  
of the display is difficult because of the large number of responses produced by several  
local oscillator harmonics generated in the mixer. To solve this problem the 8566B  
spectrum analyzer uses two methods of signal identification.  
The first is an automatic image search in which the spectrum analyzer looks for a signal  
image within a preset range of harmonic numbers. This automatic signal identification  
routine is initiated by pressing SHIFT, 3, MHz. Upon completion, one of three messages  
appears on the CRT: SIGNAL NOT IDENTIFIED, SIGNAL IDENTIFIED OUT OF BAND(with  
the frequency), or CENTER(with the identified signal frequency). The range of harmonic  
numbers through which the spectrum analyzer searches can be changed by accessing  
search start HNand search stop HN,which are SHIFT, 7, MHz and SHIFT, 8, MHz,  
respectively.  
The second method, manual signal identification, is a fast method of verifying that the  
displayed signal is a true signal in the specified band. Press SHIFT, 4, MHz to activate this  
function. Positive signal identification is indicated if the signal in question does not change  
frequency and is reduced in amplitude by one to two divisions on alternate sweeps. A failed  
signal identification is indicated by the signal disappearing or shifting in frequency.  
Chapter 2  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Operation  
Using the Mixers with the 8566B Spectrum Analyzer  
Figu r e 2-19 Sa m p le Mixer Ca libr a t ion Ta ble  
48  
Chapter2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
3 Per for m a n ce Tests  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Performance Tests  
Introduction  
In tr od u ction  
This section contains instructions for testing the performance of the 11970 Series Mixers.  
Performance tests are used to check the mixers at incoming inspection and for periodic  
evaluation. The tests verify the specifications listed for the mixers in Table 1-1.  
Test equipment required for the performance tests is listed in Table 3-1. for the 11970K,  
Table 3-2. for the 11970A, Table 3-3. for the 11970Q, Table 3-4. for the 11970U, Table 3-5. for  
the 11970V and Table 3-6. for the 11970W. Test instruments other than those listed may be  
used provided their performance equals or exceeds the critical specifications listed in  
Tables 3-1 through 3-6.  
Test Recor d  
At the end of each test is a test record, which is used for recording the performance test  
data. Make copies of these test records and use them as worksheets when doing the tests.  
50  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Performance Tests  
Performance Test Procedures  
Per for m a n ce Test P r oced u r es  
Descr ip tion  
Each performance test procedure is contained in a single paragraph. The first entry in  
each paragraph is the specification for the parameter being measured as described in Table  
1-1 This is followed by a general description of the test and any special instructions or  
problem areas. Appropriate test setup illustrations are included in this section and are  
referenced in the procedures. You MUST do the tests, and the steps within each test, in the  
order they are given.  
Table 3-1. Recommended Test Equipment for 11907K  
In st r u m en t  
Cr it ica l Sp ecifica tion s  
R ecom m en d ed Mod el  
Spectrum Analyzer  
LO and IF ranges compatible with mixer  
8566B  
8340A  
Synthesized Sweeper  
Frequency: 18 to 26.5 GHz  
Output Level: >-10 dBm  
Amplifier  
Output Level: >18 dBm leveled  
Frequency Range: 3 to 4.5 GHz  
11975A  
Power Meter  
Power Sensor  
Compatible with Power Sensor  
SWR: <1.3  
436A  
8485A  
K752C  
Directional Coupler1  
Coupling: 10 dB  
Directivity: >40 dB  
Primary Arm SWR: <1.05  
Auxiliary Arm SWR: <1.2  
Isolator  
Insertion Loss: <1.5 dB  
Isolation: >20 dB  
SWR: <1.2  
P/N 0960-0081  
K281C  
Adapter (2 required)  
3.5 mm female to WR-42  
SWR: <1.1  
Cables (3 required)  
Cable  
Connectors: SMA male  
P/N 5061-5458  
P/N 8120-4396  
Connectors: SMA  
Loss: <1.0 dB @ 20 GHz  
1. Calibration data for the coupling ratio between the output arm and the auxiliary  
arm is necessary for accurate measurements.  
Chapter 3  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Performance Tests  
Performance Test Procedures  
Table 3-2. Recommended Test Equipment for 11970A  
In st r u m en t  
Cr itica l Sp ecifica tion s  
Recom m en d ed  
Mod el  
Spectrum Analyzer  
LO and IF ranges compatible with mixer  
8566B  
8340A  
Synthesized Sweeper  
Frequency: 8 to 13.5 GHz  
Output Level: >0 dBm  
Amplifier  
Amplifier  
Output Level: >18 dBm leveled  
Frequency Range: 3 to 5 GHz  
11975A  
8349A  
Frequency Range: 8 to 13.5 GHz  
Output Level >15 dBm  
Power Meter  
Power Sensor  
Power Sensor  
Compatible with Power Sensor  
SWR: <2.0  
436A  
R8486A  
8485A  
R752D  
SWR: <1.3 @ 6 GHz  
Directional Coupler1  
Coupling: 20 dB  
Directivity: >40 dB  
Primary Arm SWR: <1.05  
Auxiliary Arm SWR: <1.2  
Isolator  
Insertion Loss: <1.5 dB  
Isolation: >20 dB  
SWR: <1.2  
P/N 0960-0082  
Spacekom Microwave2  
TKa-1  
Frequency Tripler  
Input Power: >10 dBm minimum  
Conversion Loss: <15 dB  
Adapter  
SMA female to SMA female  
SMA female to Type N female  
Type N male to SMA female  
Connectors: SMA male  
P/N 1250-1158  
P/N 1250-1404  
P/N 1250-1250  
P/N 5061-5458  
P/N 8120-4396  
Adapter  
Adapter (2 required)  
Cables (3 required)  
Cable  
Connectors: SMA male  
Loss: <1.0 dB @ 20 GHz  
Cable  
Connectors: BNC male  
11086A  
1. Calibration data for the coupling ration between the output arm and the auxil-  
iary arm is necessary for accurate measurements.  
2. Honeywell, Inc. Spacekom Microwave Center, Santa Barbara, CA  
52  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Performance Tests  
Performance Test Procedures  
Table 3-3. Recommended Test Equipment for 11970Q  
In st r u m en t  
Cr itica l Sp ecifica tion s  
R ecom m en d ed Mod el  
Spectrum Analyzer  
Synthesized Sweeper  
LO and IF ranges compatible with mixer  
8566B  
8340A  
Frequency: 11.0 to 16.7 GHz  
Output Level: >0 dBm  
Amplifier  
Amplifier  
Output Level: >18 dBm leveled  
Frequency Range: 4 to 6 GHz  
11975A  
8349A  
Frequency Range: 11.0 to 16.7 GHz  
Output Level >15 dBm  
Power Meter  
Power Sensor  
Power Sensor  
Compatible with Power Sensor  
SWR: <2.0  
436A  
Q8486A  
8485A  
Q752D  
SWR: <1.3 @ 6 GHz  
Directional Coupler1  
Coupling: 20 dB  
Directivity: >30 dB  
Auxiliary Arm SWR: <1.2  
Isolator  
Insertion Loss: <2 dB  
Isolation: >20 dB  
SWR: <1.5  
Q365A  
Spacekom Microwave2  
TKa-1  
Frequency Tripler  
Conversion Loss: <20 dB  
Adapter  
SMA female to Type N female  
SMA female to SMA female  
Type N male to SMA female  
Connectors: SMA male  
P/N 1250-1404  
P/N 1250-1158  
P/N 1250-1250  
P/N 5061-5458  
P/N 8120-4396  
Adapter  
Adapter (2 required)  
Cables (4 required)  
Cable  
Connectors: SMA male  
Loss: <1.0 dB @ 20 GHz  
Cable  
Connectors: BNC male  
11086A  
1. Calibration data for the coupling ration between the output arm and the auxil-  
iary arm is necessary for accurate measurements.  
2. Honeywell, Inc. Spacekom Microwave Center, Santa Barbara, CA  
Chapter 3  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Performance Tests  
Performance Test Procedures  
Table 3-4. Recommended Test Equipment for 11970U  
In st r u m en t  
Cr it ica l Sp ecifica tion s  
Recom m en d ed Mod el  
Spectrum Analyzer  
LO and IF ranges compatible with mixer  
8566B  
8340A  
Synthesized Sweeper  
Frequency: 13.3 to 20 GHz  
Output Level: > 9 dBm  
Amplifier  
Amplifier  
Output Level: > 18 dBm leveled  
Frequency Range: 4 to 6 GHz  
11975A  
8349A  
432A  
Frequency Range: 13.3 to 20 GHz  
Output Level >15 dBm  
Power Meter  
Compatible with Power Sensor  
SWR: <2.0  
Hughes1 45773H-1100  
Thermistor Mount  
Thermistor Mount  
SWR: <1.3 @ 6 GHz  
478A  
Directional Coupler2  
Coupling: 20 dB  
Directivity: >30 dB  
Auxiliary Arm SWR: <1.2  
U752D  
Isolator  
Insertion Loss: <2 dB  
Isolation: >20 dB  
SWR: <1.5  
U365A  
Spacekom Microwave3  
TKa-1  
Frequency Tripler  
Conversion Loss: <20 dB  
Adapter  
SMA female to Type N female  
SMA female to SMA female  
Type N male to SMA female  
Connectors: SMA male  
P/N 1250-1404  
P/N 1250-1158  
P/N 1250-1250  
P/N 5061-5458  
P/N 8120-4396  
Adapter  
Adapter (2 required)  
Cables (4 required)  
Cable  
Connectors: SMA male  
Loss: <1.0 dB @ 20 GHz  
Cable  
Connectors: BNC male  
11086A  
1. Hughes Aircraft Co. Electron Dynamics Division, Torrance, CA  
2. Calibration data for the coupling ration between the output arm and the auxil-  
iary arm is necessary for accurate measurements.  
3. Honeywell, Inc. Spacekom Microwave Center, Santa Barbara, CA  
54  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Performance Tests  
Performance Test Procedures  
Table 3-5. Recommended Test Equipment for 11970V  
In st r u m en t  
Cr itica l Sp ecifica tion s  
R ecom m en d ed Mod el  
Spectrum Analyzer  
Swept Source  
LO and IF ranges compatible with mixer 8566B  
Compatible with V-Band Plug-In  
8350B  
Hughes1 47724H-1510  
V-Band Source Plug-In  
Frequency Range: 50 to 75 GHz  
Output Level: >0 dBm  
Amplifier  
Output Level >18 dBm leveled  
Frequency Range: 4 to 6 GHz  
11975A  
Power Meter  
Compatible with Power Sensor  
SWR: <2.0  
432A  
Thermistor Mount  
Thermistor Mount  
Hughes 45774H-1100  
478A  
SWR: <1.3 @ 6 GHz  
Directional Coupler2  
Coupling: 20 dB  
Directivity: >20 dB  
Auxiliary Arm SWR: <1.5  
Hughes 45324H-1220  
TRG3 510V/385  
P/N 1250-1404  
P/N 1250-1158  
P/N 1250-1250  
P/N 5061-5458  
P/N 8120-4396  
Variable Attentuator  
Range: 0 to 15 dB  
Adapter  
SMA female to Type N female  
SMA female to SMA female  
Type N male to SMA female  
Connectors: SMA male  
Adapter  
Adapter (2 required)  
Cables (4 required)  
Cable  
Connectors: SMA male  
Loss: <1.0 dB @ 20 GHz  
Cable  
Connectors: BNC male  
11086A  
1. Hughes Aircraft Co. Electron Dynamics Division, Torrance, CA  
2. Calibration data for the coupling ration between the output arm and the auxiliary  
arm is necessary for accurate measurements.  
3. Alpha Industries Inc. TRG Division, Woburn, MA  
Chapter 3  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Performance Tests  
Performance Test Procedures  
Table 3-6. Recommended Test Equipment for 11970W  
In st r u m en t  
Cr it ica l Sp ecifica t ion s  
R ecom m en d ed  
Mod el  
Spectrum Analyzer  
Swept Source  
LO and IF ranges compatible with mixer  
Compatible with W-Band Plug-In  
8566B  
8350B  
Hughes1 47726H-1510  
W-Band Source Plug-In  
Frequency Range: 75 to 110 GHz  
Output Level: >0 dBm  
Amplifier  
Output Level >18 dBm leveled  
Frequency Range: 4 to 6 GHz  
11975A  
Power Meter  
Compatible with Power Sensor  
SWR: <2.0  
432A  
Thermistor Mount  
Thermistor Mount  
Hughes 45776H-1100  
478A  
SWR: <1.3 @ 6 GHz  
Directional Coupler2  
Coupling: 20 dB  
Directivity: >20 dB  
Auxiliary Arm SWR: <1.5  
Hughes 45326H-1220  
TRG3 W510  
Variable Attentuator  
Range: 0 to 15 dB  
Adapter  
SMA female to Type N female  
SMA female to SMA female  
Type N male to SMA female  
Connectors: SMA male  
P/N 1250-1404  
P/N 1250-1158  
P/N 1250-1250  
P/N 5061-5458  
P/N 8120-4396  
Adapter  
Adapter (2 required)  
Cables (4 required)  
Cable  
Connectors: SMA male  
Loss: <1.0 dB @ 20 GHz  
Cable  
Connectors: BNC male  
11086A  
1. Hughes Aircraft Co. Electron Dynamics Division, Torrance, CA  
2. Calibration data for the coupling ration between the output arm and the auxil-  
iary arm is necessary for accurate measurements.  
3. Alpha Industries Inc. TRG Division, Woburn, MA  
56  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Performance Tests  
Conversion Loss and Frequency Response  
Con ver sion Loss a n d Fr eq u en cy R esp on se  
Sp ecifica tion s  
Conversion Loss:  
For a CW RF input power of less than -20 dBm  
11970K: 24 dB maximum  
11970A: 26 dB maximum  
11970Q: 28 dB maximum  
11970U: 28 dB maximum  
11970V: 40 dB maximum  
11970W: 47 dB maximum  
Frequency Response:  
For an LO amplitude between 14.5 and 16.0 dBm  
11970K: ±1.9 dB  
11970A: ±1.9 dB  
11970Q: ±1.9 dB  
11970U: ±1.9 dB  
11970V: ±2.1 dB  
11970W: ±3.0 dB  
For an LO amplitude between 14.0 and 18.0 dBm  
11970K: ±2.8 dB  
11970A: ±2.8 dB  
11970Q: ±2.8 dB  
1197OU: ±2.8 dB  
11970V: ±2.8 dB  
11970W: ±4.0 dB  
Chapter 3  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Performance Tests  
Conversion Loss and Frequency Response  
Descr ip tion  
The frequency response and conversion loss are checked at four LO power levels. A known  
input power is applied to the input of the mixer. The IF output power is measured on the  
8566B Spectrum Analyzer. From these measurements, the conversion loss and frequency  
response are calculated.  
1. Connect an SMA cable from the 1st LO OUTPUT of the spectrum analyzer to the  
INPUT of the amplifier. Connect a second SMA cable to the OUTPUT of the amplifier.  
2. For 11970K: Zero and calibrate the power meter. For 11970A, 11970Q, 11970U, 11970V  
or 11970W: Connect the 478A Power Sensor to the power meter sensor cables, then zero  
the power meter.  
3. Set the 8566A/B Spectrum Analyzer controls as follows:  
SHIFT, , KSU  
FREQUENCY SPAN, 0, Hz  
SHIFT, SWEEP, CONT, KSt  
CENTER, 2, 4, [.] 5, GHz  
CAUTION  
When you are using a 11975A Amplifier with a 11970 Series Mixer, you  
MUST set the amplifier rear-panel ALC switch to ON before you connect the  
amplifier into the test setup. If the ALC switch is left in the OFF position, the  
amplifier output power is high enough to destroy the mixer diodes.  
4. On the 11975A Amplifier, set the rear panel ALC switch to ON, then connect the power  
sensor to the free end of the cable installed on the OUTPUT connector of the amplifier.  
Set the power meter Cal Factor to the appropriate value for a frequency of 4 GHz.  
5. Adjust the amplifier OUTPUT POWER LEVEL for a reading of 14.0 ± 0.1 dBm on the  
power meter. Record the LO power in Table 3-7.  
6. Set the power meter Cal Factor to 100 percent.  
7. Connect the equipment, as shown in Figure 3-1.  
CAUTION  
Make sure the 8349A Amplifier, used in the signal generator system for  
11970A, Q and U tests, is set for external leveling before you turn it on.  
Failure to do so may allow the amplifier output to rise about 20 dBm (high  
enough to damage the frequency tripler).  
8. Set the signal generator for a CW output signal at the frequencies listed below:  
11970K: 18.0 GHz  
11970A: 26.5 GHz  
11970Q: 33.0 GHz  
11970U: 40.0 GHz  
58  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Performance Tests  
Conversion Loss and Frequency Response  
11970V: 50.0 GHz  
11970W: 75.0 GHz  
Chapter 3  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Performance Tests  
Conversion Loss and Frequency Response  
9. Adjust the output power of the signal generator for a reading of approximately -10 dBm  
on the power meter for the 11970K, V or W and for approximately -3 dBm for the  
11970A, Q or U. (Make sure that the unleveled light is not on for the 11970V and  
11970W. Vary the power by adjusting the rotary vane attenuator.)  
10.On the spectrum analyzer, press SHIFT and [MKR/∆ ⇑ STP SIZE].  
11.Set spectrum analyzer CENTER FREQUENCY to:  
11970K: 22 GHz  
11970A: 33 GHz  
11970Q: 41 GHz  
11970U: 50 GHz  
11970V: 62 GHz  
11970W: 92 GHz  
12.On the spectrum analyzer, press SHIFT and SWEEP, CONT.  
13.Set the spectrum analyzer controls as follows:  
For 11970K, A, Q or U:  
SHIFT, REFERENCE LEVEL, 3, 0, -dB  
SCALE ENTER dB/DIV, 2, dB  
RES BW, 1, MHz  
FREQUENCY SPAN, 2, 0, MHz  
CF STEP SIZE, 5, 0, 0, MHz  
For 11970V or W:  
SHIFT, REFERENCE LEVEL, 3, 0, -dB  
SCALE ENTER dB/DIV, 2, dB  
RES BW, 3, MHz  
FREQUENCY SPAN, 2, 0, 0, MHz  
CF STEP SIZE, 1, GHz  
Press REFERENCE LEVEL:  
11970K: 3, 6, -dBm  
11970A: 3, 8, -dBm  
11970Q: 4, 0, -dBm  
11970U: 4, 0, -dBm  
11970V: 4, 8, -dBm  
11970W: 5, 6, -dBm  
60  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Performance Tests  
Conversion Loss and Frequency Response  
Then press CENTER FREQUENCY:  
11970K: 1, 8, GHz  
11970A: 2, 6, [.] 5, GHz  
11970Q: 3, 3, GHz  
11970U: 4, 0, GHz  
11970V: 5, 0, GHz  
11970W: 7, 5, GHz  
14.Press MARKER, PEAK SEARCH. If necessary, press MKRCF and use DISPLAY LINE  
ENTER to find the average of the signals peak variations. (When testing 11970V or  
11970W mixers, it is important to re-zero the power meter for each measurement.)  
15.Record the following in Table 3-7:  
Marker Frequency  
Marker Amplitude  
Power Meter Reading  
Power Sensor Cal Factor or Correction Factor (dB)  
Directional Coupler Coupling Factor  
NOTE  
For the purposes of this measurement, the directional coupler coupling factor  
is defined as the ratio of the power at the output flange to the power at the  
coupled flange.  
16.Calculate the conversion loss of the mixer with the following equation:  
Conversion Loss = Power Meter Reading - 10 log(Cal Factor) - Spectrum Analyzer  
Marker Amplitude - Coupling Factor  
(or given the Power Meter Correction Factor in dB: Conversion Loss = Power Meter  
Reading + Power Meter Correction Factor - Spectrum Analyzer Marker Amplitude -  
Coupling Factor)  
For example:  
Power Meter Reading = -10.03 dBm  
Cal Factor = 94.8%  
or Correction Factor = -0.232 dB  
Spectrum Analyzer reading = -39.78 dBm  
Coupling Factor 8.93 dB  
then:  
Conversion Loss (-10.03) - 10 log(.948) - (-39.78 dBm) - 8.93 dB = 21.05 dB  
Chapter 3  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Performance Tests  
Conversion Loss and Frequency Response  
or Conversion Loss = (-10.03 dBm) - (-0.232) - (-39.78 dBm) - 8.93 dB = 21.05 dB  
17.Record the conversion loss in Table 3-7.  
62  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Performance Tests  
Conversion Loss and Frequency Response  
NOTE  
The conversion loss indicated on the mixer calibration label includes the loss  
in the IF cable. If other than the specified cable is used, then the loss in that  
cable must be compensated for when making amplitude measurements.  
18.Increment the frequency of the signal generator 500 MHz higher.  
19.Press DATA [], then PEAK SEARCH and MKRCF on the spectrum analyzer.  
20.Repeat steps 15 through 18 until the appropriate frequency listed below is reached.  
11970K: 26.5 GHz  
11970A: 40.0 GHz  
11970Q: 50.0 GHz  
1197OU: 60.0 GHz  
11970V: 75.0 GHz  
11970W: 110.0 GHz  
21.Repeat steps 1 through 19 for LO inputs to the mixer of 14.5 dBm, 16.0 dBm, and  
18.0 dBm. In step 5, measure each of these levels at the end of the cable normally  
connected to the mixer LO input.  
22.Frequency response is the difference between the maximum and minimum conversion  
losses recorded in Table 3-7. For LO power levels between 14.5 and 16.0 dBm this  
difference must be less than:  
3.8 dB (for 11970K, 11970A, 11970Q or 1197OU)  
4.2 dB (for 11970V)  
6.0 dB (for 11970W)  
For LO power levels between 14.0 and 18.0 dBm, the difference must be less than:  
5.6 dB (for 11970K, 11970A, 11970Q, 11970U or 11970V)  
8.0 dB (for 11970W)  
23.Maximum conversion loss must not exceed the following limits:  
For an LO input power between 14.0 and 18.0 dBm.  
11970K: 24 dB  
11970A: 26 dB  
11970Q: 28 dB  
11970U: 28 dB  
11970V: 40 dB  
11970W: 46 dB  
Chapter 3  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Performance Tests  
Conversion Loss and Frequency Response  
Figu r e 3-1 Per for m a n ce Test Set u p s (1 of 2)  
64  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Performance Tests  
Conversion Loss and Frequency Response  
Figu r e 3-2 Per for m a n ce Test Set u p s (2 of 2)  
Chapter 3  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Performance Tests  
Conversion Loss and Frequency Response  
Table 3-7 Conversion Loss and Frequency Response Test Record  
CONVER SION LOSS AND F R E QUENCY R ESP ONSE for a n LO P OWER  
of_____________d Bm  
Model Number____________ Date_____________  
Serial Number______________Tested By _________________  
Marker  
Frequency  
Marker  
Amplitude  
Power Meter  
Readings  
Power Sensor  
Cal Factor  
Directional Coupler  
Coupling Factor  
Conversion  
Loss  
GHz  
dBm  
dBm  
% or dB  
dB  
dB  
66  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Performance Tests  
Conversion Loss and Frequency Response  
Table 3-7 Conversion Loss and Frequency Response Test Record  
CONVER SION LOSS AND F R E QUENCY R ESP ONSE for a n LO P OWER  
of_____________d Bm  
Marker  
Frequency  
Marker  
Amplitude  
Power Meter  
Readings  
Power Sensor  
Cal Factor  
Directional Coupler  
Coupling Factor  
Conversion  
Loss  
GHz  
dBm  
dBm  
% or dB  
dB  
dB  
Frequency Response = _______________dB  
Chapter 3  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Performance Tests  
AVERAGE NOISE LEVEL TEST  
AVERAGE NOISE LEVEL TEST  
Sp ecifica tion  
11970K: -110 dBm  
11970A: -108 dBm  
11970Q: -104 dBm  
11970U: -104 dBm  
11970V: -92 dBm  
11970W: -85 dBm  
Descr ip tion  
The average displayed noise level in a 1 kHz bandwidth, using external mixing with the  
8566A/B Spectrum Analyzer, is checked at several LO power levels. This is accomplished  
by applying a known power to the input of the mixer. The difference between the  
amplitude of the known signal and the noise floor is measured. From these measurements,  
the average noise level in a 1 kHz bandwidth is calculated.  
1. Connect an SMA cable from the spectrum analyzer LO OUTPUT to the INPUT of the  
11975A Amplifier. Connect a second SMA cable to the amplifier OUTPUT.  
2. For 11970K: Zero and calibrate the power meter. For 11970A, 11970Q, 11970U, 11970V  
or 11970W: Connect the 478A Power Sensor to the power meter sensor cables, then zero  
the power meter.  
3. On the 8566A/B Spectrum Analyzer, set the controls as follows:  
SHIFT, DATA STEP [] KSU  
FREQUENCY SPAN, 0, Hz  
SHIFT, SWEEP, CONT, KSt  
CENTER FREQUENCY, 2, 4 [.] 5, GHz  
CAUTION  
When you are using a 11975A Amplifier with a 11970 Series Mixer, you  
MUST set the amplifier rear-panel ALC switch to ON before you connect the  
amplifier into the test setup. If the ALC switch is left in the OFF position, the  
amplifier output power is high enough to destroy the mixer diodes.  
68  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Performance Tests  
AVERAGE NOISE LEVEL TEST  
4. On the 11975A Amplifier, set the rear panel ALC switch to ON. Then connect the power sensor to  
the free end of the cable connected to the output of the amplifier. Set the power meter calibration  
factor to the value shown on the power sensor calibration label for 4 GHz.  
5. Adjust the amplifier OUTPUT POWER LEVEL control for a reading of 14.5 ± 0.1 dBm  
on the power meter.  
Chapter 3  
69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Performance Tests  
AVERAGE NOISE LEVEL TEST  
CAUTION  
Make sure the 8349A Amplifier, used in the signal generator system for  
11970A, Q and U tests, is set for external leveling before you turn it on.  
Failure to set this amplifier for external leveling may allow the amplifier  
output to rise about 20 dBm, which is high enough to damage the frequency  
tripler.  
6. Disconnect the power sensor and connect the equipment as shown in Figure 3-1.  
7. On the 8566A/B Spectrum Analyzer, set the controls as follows:  
For 11970K, A, Q or U:  
SHIFT, DATA STEP [] KSU  
FREQUENCY SPAN 3, 0, MHz  
VIDEO BW, 1, 0, 0, Hz  
RES BW, 1, 0, 0, kHz  
For 11970V or W:  
SHIFT, DATA STEP [] KSU  
FREQUENCY SPAN, 1, 0, 0, MHz  
VIDEO BW, 3, MHz  
RES BW, 3, MHz  
and CENTER FREQUENCY:  
For 11970K: 2, 2, GHz  
For 11970A: 3, 3, GHz  
For 11970Q: 4, 1, GHz  
For 11970U: 5, 0,GHz  
For 11970V: 6, 2,[.] 5, GHz  
For 11970W: 9, 2,[.] 5,GHz  
8. Press SHIFT and SWEEP, CONT, KSt.  
9. Press CENTER FREQUENCY again and enter the appropriate center frequency as follows:  
For 11970K: 1, 8, GHz  
For 11970A: 2, 6, [.] 5, GHz  
For 11970Q: 3, 3, GHz  
For 11970U: 4, 0, GHz  
70  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Performance Tests  
AVERAGE NOISE LEVEL TEST  
For 11970V: 5, 0, GHz  
For 11970W: 7, 5, GHz  
10.Record the center frequency in Table 3-8.  
11.Set the power meter CAL FACTOR to the value shown on the sensor calibration label  
for the center frequency indicated on the spectrum analyzer.  
12.Set the signal generator output to produce a CW signal near the center of the spectrum  
analyzer display. Next, set the signal generator output power level to produce a reading  
of approximately -10 dBm on the power meter. Record the power meter reading in Table  
3-8. (When testing the 11970V or W, re-zero the power meter for each measurement.)  
13.Subtract the coupling factor for the directional coupler from the power meter reading,  
then add the power meter correction factor (dB) to the power meter reading. Record this  
corrected power reading in Table 3-8.  
Corrected Power Reading = Power Meter Reading - Coupling Factor + Power Meter  
Correction Factor.  
For example:  
(-10.03 dBm) - 9.82 dB + 0.73 = -19.1 dBm  
14.On the 8566A/B, press MARKER, PEAK SEARCH, then MKRREF LVL. Record the marker  
amplitude in Table 3-8.  
15.Move the marker to the displayed noise floor, or turn off the CW signal, and record the  
marker amplitude in Table 3-8. (Some sources may have excessive wide band noise when  
the RF signal is turned off, so that the noise floor must be measured with the signal left  
on.)  
NOTE  
Step 15 requires a signal source with a wide band phase noise characteristic  
that is at least 6 dB better than the measured average noise level.  
16.The Average Noise Level is calculated as follows:  
Average Noise Level = Corrected Power Reading (step 13) - Marker Amplitude (step 14)  
+ Marker Amplitude (step 15) - Measurement Bandwidth Correction Factor  
11970K, A, Q and U Bandwidth Correction = 10log(l00kHz/ 1 kHz)  
11970V and W Bandwidth Correction = 10log(3MHz/ 1 kHz)  
For example:  
Average Noise Level = (-19.1 dBm) - (-32.3 dBm) + (-77.7 dBm) - 34.77 dB = - 99.3  
dBm  
17.Enter the calculated value in Table 3-8.  
The Average Noise Level must be less than:  
-110 dBm (for the 11970K)  
-108 dBm (for the 11970A)  
Chapter 3  
71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Performance Tests  
AVERAGE NOISE LEVEL TEST  
-104 dBm (for the 11970Q)  
-104 dBm (for the 11970U)  
-92 dBm (for the 11970V)  
-85 dBm (for the 11970W)  
18.Enter the following push button commands on the 8566A/B:  
MARKER, OFF  
REFERENCE LEVEL, 0, dBm  
19.Repeat steps 9 through 17 at the following center frequencies:  
11970K: 22 and 26.5 GHz  
11970A: 33 and 40 GHz  
11970Q: 41 and 50 GHz  
11970U: 50 and 60 GHz  
11970V: 62 and 75 GHz  
11970W: 92 and 110 GHz  
20.Repeat steps 1 through 18 for a power level of 16.0 dBm at the 11970 Mixer LO  
connector.  
Table 3-8 Average Noise Test Record  
AVE R AGE NOISE LE VEL  
Model Number______________ Date______________  
Serial Number________________ Tested By______________  
Center  
Frequency  
Power  
Meter  
Reading  
Power  
Meter  
Correction  
Factor  
Directional  
Coupler  
Coupling  
Factor  
Corrected  
Power  
REading  
Signal  
Marker  
Amplitude  
Noise  
Floor  
Marker  
Amplitude  
Bandwidth  
Correction  
Factor  
Average  
Noise  
Level  
Units  
Step  
GHz  
10  
dBm  
12  
dB  
13  
dB  
13  
dBm  
13  
dBm  
14  
dB  
15  
dB  
16  
dBm  
16  
LO  
Power  
= 14.5  
LO  
Power  
= 16.0  
72  
Chapter3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
4 Ser vice  
73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Maintenance  
Ma in ten a n ce  
The only maintenance required for the 11970 Series Mixers is preventive maintenance.  
When you are not using your mixer, cover its waveguide input with its waveguide cap.  
Also, though the 11970 Mixers can absorb more punishment than is normal for such  
devices, you should avoid subjecting them to unnecessary shock or vibration.  
Rep a ir s  
The 11970 Mixers are NOT field-repairable. If your mixer fails, DO NOT try to repair it  
yourself, you will void the warranty. Instead, notify the nearest Agilent office.  
Rep la cea ble Pa r ts  
For a list of replaceable parts and accessories, see Table 4-2.  
Cir cu it Descr ip tion  
A schematic diagram of a 11970 Series Harmonic Mixer is shown in Figure 4-1. The mixer  
circuit employs two diodes arranged as an anti-parallel pair. These diodes are the  
termination for the open end of the waveguide output. By employing a matched diode pair,  
even harmonic mixing is enhanced while odd harmonic mixing is suppressed.  
The waveguide input is exponentially tapered in both height and width. The height taper  
provides impedance matching between the high impedance waveguide input and the low,  
dynamic impedance of the diodes. The width taper creates a high-pass filter response  
which isolates the LO harmonics from the standard-height waveguide. Without this  
isolation, the LO harmonics would reflect from the standard-height waveguide back into  
the mixer and destructively interfere with the desired mixing product.  
LO harmonics are confined to the immediate vicinity of the diode pair by the low-pass  
filter, which has as its first element a metal -insulator -semiconductor (MIS) capacitor.  
This improves the out-of-band response. The diplexer separates the 3 - 6 GHz LO signal  
from the 321.4 MHz IF signal.  
74  
Chapter4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Service  
Maintenance  
Figu r e 4-1 11970 Ser ies Mixer Sch em a t ic Dia gr a m  
Table 4-1  
LP F F co  
Mod el  
(GHz)  
11970K  
11970A  
11970Q  
11970U  
11970V  
11970W  
4.4  
5.0  
5.0  
6.2  
5.4  
6.2  
Chapter 4  
75  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Maintenance  
Table 4-2 Accessories and Replaceable Parts  
Pa r t Nu m ber  
Descr ip t ion  
5061-5460:  
Mixer Connector Kit (Option 009), includes the following three  
items:  
5061-5458  
8710-0510  
8710-1539  
Cable, 1 meter long, SMA male connectors (3 required)  
Wrench, 5/16-inch, open-end  
Ball Driver, 3/32-inch  
3030-0221  
1390-0671  
Socket Head Cap Screw, 4-40 thread, .375 inches long (flange  
connecting screw for 11970K and 11970A)  
Socket Head Cap Screw, captive, 4-40 thread, .290 inches long  
(flange connecting screw for 11970Q, 11970U, 11970V or  
11970W)  
11970-40001  
11970-40002  
11970-40003  
08486-40103  
11970K Waveguide Cap  
11970A Waveguide Cap  
Waveguide Cap for 11970Q or 11970U  
Waveguide Cap for 11970V or 11970W  
76  
Chapter4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Maintenance  
Table 4-3 Agilent Technologies Sales and Service Offices  
UNITED STATES  
Instrument Support Center  
Agilent Technologies  
(800) 403-0801  
EUROP EAN F IELD OP ERATIONS  
Headquarters  
France  
Germany  
Agilent Technologies S.A.  
150, Route du Nant-dAvril  
1217 Meyrin 2/ Geneva  
Switzerland  
Agilent Technologies France  
1 Avenue Du Canada  
Zone DActivite De  
Courtaboeuf  
Agilent Technologies GmbH  
Agilent Technologies Strasse  
61352 Bad Homburg v.d.H  
Germany  
(41 22) 780.8111  
F-91947 Les Ulis Cedex  
France  
(49 6172) 16-0  
(33 1) 69 82 60 60  
Great Britain  
Agilent Technologies Ltd.  
Eskdale Road, Winnersh  
Triangle Wokingham, Berkshire  
RG41 5DZ England  
(44 118) 9696622  
INTERCON F IELD OP ERATIONS  
Headquarters  
Australia  
Canada  
Agilent Technologies  
3495 Deer Creek Rd.  
Palo Alto, CA 94304-1316  
USA  
Agilent Technologies Australia  
Ltd.  
31-41 Joseph Street  
Blackburn, Victoria 3130  
(61 3) 895-2895  
Agilent Technologies (Canada)  
Ltd.  
17500 South Service Road  
Trans-Canada Highway  
Kirkland, Quebec H9J 2X8  
Canada  
(415) 857-5027  
(514) 697-4232  
Japan  
Agilent Technologies Japan, Ltd.  
Measurement Assistance Center  
9-1, Takakura-Cho, Hachioji-Shi, Singapore (Pte.) Ltd.  
Tokyo 192-8510, Japan  
TEL (81) -426-56-7832  
FAX (81) -426-56-7840  
Singapore  
Agilent Technologies  
Taiwan  
Agilent Technologies Taiwan  
8th Floor, H-P Building  
337 Fu Hsing North Road  
Taipei, Taiwan  
150 Beach Road  
#29-00 Gateway West  
Singapore 0718  
(65) 291-9088  
(886 2) 712-0404  
China  
China Agilent Technologies  
38 Bei San Huan X1 Road  
Shuang Yu Shu  
Hai Dian District  
Beijing, China  
(86 1) 256-6888  
Chapter 4  
77  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Service  
Maintenance  
78  
Chapter4  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Accell Computer Hardware MPC 68K Series User Manual
Airgle Air Cleaner AG800 User Manual
AKG Acoustics Headphones K142HD User Manual
AKG Acoustics Microphone C 414 User Manual
Alesis Musical Instrument K2661 User Manual
American Dryer Corp Clothes Dryer AD 200 Tilting User Manual
ATN Binoculars Night Vision Biocular User Manual
Avenview TV Converter Box HDMI 3G HD SD SDI User Manual
Badger Basket Camping Equipment 08100 User Manual
Beko Hot Beverage Maker BKK2213 User Manual