Agilent Technologies G3180B User Manual

Agilent G3180B  
Two-Way Splitter Kit  
With Makeup Gas  
Installation and Operation  
Guide  
Agilent Technologies  
In this Guide. . .  
This Installation and Operation Guide contains information for installing and  
using an effluent splitter on an Agilent 6890 gas chromatograph (GC). The  
G3180 splitter is intended for use with capillary columns and uses makeup gas  
to maintain adequate flows throughout the system.  
1
2
3
Introduction  
This chapter describes how the splitter works, the GC and software  
requirements of the system and the contents of the installation kit.  
Hardware Installation  
See this chapter for a detailed procedure for installing the splitter hardware  
and connecting the makeup gas supply.  
Splitter Configurations  
The split ratio (how the column effluent divides between the two detectors) is  
governed by two restrictors, which are lengths of deactivated fused silica  
tubing. This chapter presents a set of precalculated “typical” configurations. If  
desired, you can create a custom configuration to meet specific needs. The  
chapter describes a set of software tools, included in the kit, to assist you in  
designing such configurations. Finally, installation of the column and  
restrictors is covered.  
4
Operation  
This chapter contains a worked-out custom configuration, plus a few special  
topics.  
Installation and Operation Guide  
3
4
Installation and Operation Guide  
Contents  
1
Introduction  
Overview  
How It Works  
Details 10  
Metal ferrules 10  
Calculation of chromatographic parameters 11  
GC Requirements 12  
Other Requirements 12  
Parts Supplied 13  
2
Hardware Installation  
Install the Bracket and Splitter 21  
Connect the Makeup Gas Supply 24  
To supply the makeup gas froma PCM 24  
To supply the makeup gas from an Auxiliary Pressure  
controller 24  
Installation and Operation Guide  
5
3
Splitter Configurations  
Custom Configurations 29  
Restrictor id and length 32  
Inlet pressure 34  
4
Operation  
An Example 40  
Column flow 40  
Select restrictors 42  
Calculate ECD restrictor flow 44  
Calculate MSD restrictor flow 45  
Changing Columns Without Venting the MSD 46  
Backflushing the Column 47  
6
Installation and Operation Guide  
Agilent G3180B Splitter Kit  
Installation and Operation Guide  
1
Overview  
How It Works  
Details 10  
Other Requirements 12  
Parts Supplied 13  
Part Identification 14  
Parts Not Supplied 15  
Tools Required 15  
This manual covers the installation and operation of the G3180B effluent  
splitter with makeup gas kit on the Agilent 6890 series gas chromatograph  
(GC).  
7
Agilent Technologies  
 
1
Introduction  
Overview  
Splitter installation is done in three steps:  
1 Hardware installation. This gets the hardware installed and the gas flows  
connected.  
2 Restrictor configuration. You can choose to use a typical, precalculated  
configuration or create a custom one using software tools supplied on a CD.  
3 Restrictor and column installation. Using the results of step 2, cut the  
appropriate lengths of the appropriate diameter tubing for the restrictors.  
Install the restrictors and the analytical column.  
8
Installation and Operation Guide  
   
Introduction  
1
How It Works  
The splitter divides the effluent from a column between two different  
detectors. The detectors can be operating at different pressures, that is, any  
mix of the following can be used:  
• Atmospheric pressure  
FID (flame ionization detector)  
TCD (thermal conductivity detector)  
NPD (nitrogen phosphorus detector)  
ECD (electron capture detector)  
FPD (flame photometric detector)  
• Below atmospheric pressure  
MSD (mass selective detector)  
• Above atmospheric pressure  
AED (atomic emission detector)  
The split ratio is determined by the length and diameter of tubing connecting  
the splitter to the detectors. Tubing dimensions may be determined from  
Table 2 on page 26 in this manual or from a spreadsheet calculator that is  
included for calculating tubing dimensions for special situations.  
Figure 1 shows the plumbing configuration for the G3180B splitter.  
Inlet  
Det 1  
Det 2  
Restrictor 1  
Restrictor 2  
Splitter  
Constant pressure  
makeup gas supply  
Column  
GC oven  
Figure 1  
Splitter plumbing  
Installation and Operation Guide  
9
             
1
Introduction  
The column flow mixes with the makeup flow in the splitter. This mixture then  
flows through lengths of uncoated, deactivated, fused-silica tubing to each  
detector. These tubes act as flow restrictors. While the flow through each  
restrictor changes with oven temperature, the ratio of the two flows at any  
temperature is the same.  
Details  
The G3180B kit addresses several limitations of previous approaches to  
splitting column effluent between two detectors:  
Metal ferrules  
The splitter uses metal column ferrules, which eliminate air leakage into the  
sample stream. Unlike polyimide, metal ferrules do not loosen upon thermal  
cycling of the oven. They also do not outgas contaminants or shed particles  
(like graphite) that can result in chromatographic problems.  
Microfluidic plate  
The splitting hardware is based on microfluidic plate technology. This allows  
very low dead volume connections between the column end and the two  
detector restrictor tubes. The thin metal plate has fast thermal response and is  
mounted solidly on the oven wall for ease of use. The interior plate surfaces  
are deactivated to prevent adsorption by active compounds.  
Constant pressure operation  
The splitter uses a source of makeup gas supplied by electronic pneumatics  
control (EPC). This maintains the splitter at a known and constant pressure.  
Constant pressure allows easier splitting to vacuum detectors like the MSD. It  
simplifies choice of splitter parameters, allowing all aspects of the  
chromatographic setup to be calculated. Constant pressure makeup allows the  
column to be run in constant flow mode while still maintaining a constant  
split ratio between two detectors of different operating pressures such as the  
FPD and the MSD. Because the EPC pressure can be time programmed, useful  
operations like backflushing unwanted heavy materials from the column and  
changing columns in MSD systems without venting are possible.  
10  
Installation and Operation Guide  
                     
Introduction  
1
Calculation of chromatographic parameters  
Because the pressure at the split point is known and constant, the  
chromatographic parameters can be calculated before setup. This is especially  
useful with GC/MSD setups, where there are limitations on the flow rates of  
carrier gas allowed into the MSD. If a method that was originally developed on  
an MSD is converted to a splitter setup, a new inlet pressure can be calculated  
to produce retention times very similar to the original method.  
Installation and Operation Guide  
11  
   
1
Introduction  
GC Requirements  
The splitter mounts in an Agilent 6890 series GC.  
The splitter requires an electronically controlled pressure source such as the  
Three Channel Pressure controller (6890 option 205, 301, or 308) or a  
Pneumatics Control Module (PCM).  
Other Requirements  
The calculator requires Microsof®t Excel 97 (or later), which is not supplied  
with this kit.  
12  
Installation and Operation Guide  
             
Introduction  
1
Parts Supplied  
The G3180B kit contains the following parts (Table 1).  
Table 1  
Parts supplied  
Description  
Part number  
0100-0124  
Quantity  
Union, stainless steel, 1/16-inch tubing  
Union, stainless steel, 1/8 to 1/16-inch reducing  
Valve box blanking plate  
2
1
1
4
7
1
2
1
0100-0241  
G1580-00130  
G1530-01340  
0515-0374  
Capillary column spring clips  
Screw, M3 × 10 mm  
G2855-60140  
G2855-60560  
G2855-80022  
G3180-90120  
0100-2354  
Oven bracket assembly  
T-screw oven bracket retainer  
Manual and calculator CD  
Manual, G3180B  
Tubing, stainless steel, 1/16-inch od × 0.01-inch id, 1 m  
Compact splitter with makeup gas assembly  
Supplies and spares kit  
1
1
1
G3180-61500  
G2855-60150  
Installation and Operation Guide  
13  
     
1
Introduction  
Part Identification  
Most of the kit parts are easily recognized. The unique ones are identified in  
Figure 2.  
Capillary column  
spring clips  
Valve box  
blanking plate  
Oven bracket  
assembly  
This assembly is shipped in a  
plastic bag to keep contaminants  
out of the tubing and the fittings.  
Do not open the bag until you are  
ready to install the splitter.  
Compact splitter  
with makeup gas  
assembly  
Figure 2  
Part identification  
14  
Installation and Operation Guide  
   
Introduction  
1
Parts Not Supplied  
Brown-dot frit (19231-60610)  
Tools Required  
Side cutter, large  
Open-end wrenches  
Installation and Operation Guide  
15  
   
1
Introduction  
16  
Installation and Operation Guide  
Agilent G3180B Splitter Kit  
Installation and Operation Guide  
2
Prepare the GC 18  
Install the Column Clips 20  
Install the Bracket and Splitter 21  
Connect the Makeup Gas Supply 24  
This chapter describes the procedure for installing the splitter hardware and  
connecting the makeup gas supply.  
17  
Agilent Technologies  
 
2
Hardware Installation  
Prepare the GC  
Turn the power off and disconnect the power cord before proceeding.  
WARNING  
1 Raise the GC top cover to expose the oven top.  
2 Remove the valve box cutout using a side cutter (Figure 3).  
Cut tabs  
around edges  
and remove  
Figure 3  
Remove the valve box cutout  
18  
Installation and Operation Guide  
   
Hardware Installation  
2
3 This exposes a layer of soft insulation. Remove it to expose the hard oven  
insulation. Remove the precut insulation piece at the location shown in  
Figure 4.  
Remove  
this  
cutout  
Figure 4  
Remove the insulation cutout  
4 Replace the soft insulation. Install the valve box blanking plate, using one  
screw at the front and one at the rear to secure it. See Figure 5.  
Two screws  
Figure 5  
Install valve box blanking plate  
Installation and Operation Guide  
19  
   
2
Hardware Installation  
Install the Column Clips  
Install the four column clips on the oven shroud (Figure 6).  
Clip  
(1 of 4)  
Oven  
shroud  
Figure 6  
Oven clips  
20  
Installation and Operation Guide  
   
Hardware Installation  
2
Install the Bracket and Splitter  
The splitter is usually installed on the right side of the oven.  
The body of the splitter may be discolored as a result of the deactivation process. This is  
not a defect.  
NOTE  
1 Place the bracket against the side of the oven. The two notches should be up  
and the standoffs should face the center of the oven.  
2 Use two T-shaped thumbscrews to fasten the bracket to the T-slots in the  
oven wall (Figure 7).  
Thumbscrews  
Bracket  
Detail  
Figure 7  
Installing the bracket  
Installation and Operation Guide  
21  
   
2
Hardware Installation  
Use extreme care to prevent any fragments of insulation or other material from  
entering the makeup gas tubing or the fittings on the splitter assembly. Such materials  
could block the internal passages in the splitter or the bore of the capillary restrictors.  
CAUTION  
CAUTION  
In the following steps, bend the tubing over an object such as your thumb to avoid  
kinks.  
3 Open the plastic bag and remove the splitter assembly. Install a plastic cap  
on the end of the makeup gas tubing. Place small pieces of tape over the  
open end of the fittings.  
4 Prebend the tubing according to Figure 8. This will make splitter  
installation much easier.  
Up through hole  
in oven top and  
blanking plate  
60 mm  
195 mm  
Oven side wall  
140 mm  
Dimensions are approximate  
All bends are 90°  
Splitter assembly  
45 mm  
Figure 8  
Prebending the splitter tubing  
5 Push the end of the makeup gas tubing up through the top oven wall so that  
the end of the tubing comes out in the hole of the valve box blanking plate.  
22  
Installation and Operation Guide  
 
Hardware Installation  
6 Route the prebent tubing against the oven wall and top to keep it clean for  
future maintenance. It should be behind the back detector location.  
7 Screw the splitter assembly to the bracket (three screws). See Figure 9.  
Makeup gas tubing  
Mounting screws  
Splitter assembly  
Ferrule release hole  
Mounting screw  
Figure 9  
Installing the splitter assembly  
Installation and Operation Guide  
23  
 
2
Hardware Installation  
Connect the Makeup Gas Supply  
Connect the makeup gas source to the PCM or Auxiliary Pressure controller.  
To supply the makeup gas froma PCM  
1 Connect the tubing from the PCM to the 1 meter length of stainless steel  
tubing from the kit with a union.  
2 Connect the free end of the stainless steel tubing to the tubing from the  
splitter assembly with a union. See Figure 10.  
Stainless  
steel  
tubing  
To splitter  
PCM  
Union  
Union  
Figure 10 Plumbing a PCM makeup supply  
To supply the makeup gas from an Auxiliary Pressure controller  
1 Install the brown-dot frit (part no. 19231-60610) in the output channel. See  
your GC manual for details.  
2 Connect the tubing from the Auxiliary Pressure controller to the tubing  
from the splitter assembly with the 1/8 to 1/16-inch stainless steel reducing  
union.  
This completes the hardware installation.  
24  
Installation and Operation Guide  
           
Agilent G3180B Splitter Kit  
Installation and Operation Guide  
3
Typical Configurations 26  
Inlet pressure 34  
Restrictor and Column Installation 35  
Install the column 35  
Connect the splitter 35  
Disconnect tubing from the splitter 37  
The combination of restrictor diameters and lengths determines how the  
column effluent is divided (the split ratio) between the two detectors. There  
are two approaches to setting up a splitter method.  
Use a typical configuration. A set of eight configurations is discussed  
beginning on the next page. They apply to a variety of detector  
combinations and split ratios. All of the flows have been calculated.  
Create a custom configuration. If the typical configurations do not meet  
your needs, you can create one that does. The CD shipped with the splitter  
kit provides tools for the necessary calculations.  
We suggest examining the typical configurations first, since they cover a wide  
variety of splitter applications and require no calculations.  
25  
Agilent Technologies  
     
3
Splitter Configurations  
Typical Configurations  
The important parameters when setting up a splitter are the lengths and  
diameters of the restrictor tubes that go to the two detectors. The dimensions  
of the restrictors are chosen to give the desired split ratio, flow to the detector,  
and to minimize peak broadening.  
The splitter restrictors are chosen based on:  
The range of column flows that will be used with the method  
The operating pressure of the two detectors  
The flow rate requirements of the two detectors  
Table 2 lists typical splitting configurations. Table 3 shows the resulting gas  
flows. All calculations assume helium as the carrier gas.  
Table 2  
Restrictor configurations  
Configuration  
Det 1  
Det 2  
Split ratio,  
Det 2/Det 1  
Diam R1,  
mm id  
Length R1,  
m
Diam R2,  
mm id  
Length R2,  
m
1
2
3
4
5
6
7
8
atm*  
atm  
atm  
atm  
atm  
atm  
atm  
atm  
atm  
atm  
1
5
1
1
2
2
5
5
0.25  
0.18  
0.18  
0.18  
0.18  
0.18  
0.10  
0.18  
0.544  
0.418  
1.060  
0.530  
2.130  
1.064  
0.507  
2.660  
0.25  
0.25  
0.18  
0.18  
0.18  
0.18  
0.18  
0.18  
0.544  
0.311  
2.890  
1.440  
2.890  
1.443  
2.890  
1.443  
MSD,D**  
MSD, T***  
MSD, D  
MSD, T  
MSD, D  
MSD, T  
* atm Atmospheric pressure detectors such as FID, TCD, ECD, FPD and NPD  
** MSD, D MSD with diffusion pump or standard turbo pump (2 mL/min flow capability)  
*** MSD, T MSD with performance turbo pump (4 mL/min flow capability); makeup pressure supply is set to 3.8 psig  
26  
Installation and Operation Guide  
                           
Splitter Configurations  
3
Table 3  
Splitter flows  
40 °C  
200 °C  
Flow R1,  
300 °C  
Flow R1, Flow R2,  
400 °C  
Flow R2,  
Configuration Flow R1,  
mL/min  
Flow R2,  
mL/min  
Flow R2,  
mL/min  
Flow R1,  
mL/min  
mL/min  
mL/min  
mL/min  
mL/min  
1
2
3
4
5
6
7
8
14.7  
5.1  
2
14.7  
7.3  
2.5  
1
7.3  
12.7  
1
5.2  
5.2  
3.9  
1.4  
3.9  
25.6  
2
1.8  
9.1  
6.8  
0.7  
0.7  
0.54  
1.1  
0.54  
1.1  
4
4
2
2
1.4  
1.4  
1
2
0.5  
1
1
0.36  
0.71  
0.14  
0.28  
0.72  
1.4  
0.27  
0.53  
0.1  
0.54  
1.06  
0.5  
2
4
2
0.4  
0.8  
2
0.2  
0.4  
1
0.7  
4
2
1.4  
0.21  
1.1  
To use the tables, select the configuration you wish to set up. For example,  
Configuration 1 splits column effluent equally between two atmospheric  
pressure detectors (FID, TCD, ECD, FPD, and NPD). To plumb this system,  
0.544-m lengths of 0.25-mm id uncoated deactivated fused silica tubing are  
connected as restrictors from the splitter to the two detectors.  
The makeup supply (either Aux EPC or PCM module) is set to 3.8 psig. This  
will add sufficient makeup flow to the column flow to maintain the splitter  
(and thus the column outlet) at 3.8 psi. Column flow can be varied from 0 to a  
maximum flow which is determined by the upper temperature of the GC oven  
program.  
If Configuration 1 is used with a method that programs to 200 °C using helium,  
the flow through each restrictor at 200 °C will be 7.3 mL/min. The total flow  
will be 14.6 mL/min. The maximum column flow should be equal to the total  
flow minus about 1 mL/min to ensure that there is some flow for the makeup  
supply to regulate with.  
The column flow at 200 °C should be no more than 13.6 mL/min. This becomes  
important when the column is run in constant flow mode. If constant flow  
mode is used with Configuration 1 and the method programmed to 400 °C, the  
column flow should not exceed 6.8 mL/min ([3.9 + 3.9] –1).  
Installation and Operation Guide  
27  
                   
3
Splitter Configurations  
For constant pressure methods, first find the maximum flow as above. Use the  
GC, ChemStation, Flow Calculator Software or the Method Translation  
Software to find the inlet pressure that gives the maximum flow at the upper  
temperature of the method (make sure the column outlet pressure is set to  
3.8 psig for the calculation).  
helium carrier and programming to 300 °C, the pressure that gives a flow of  
9.4 mL/min ([5.2 + 5.2] – 1 = 9.4) is 56.3 psig. This is the maximum pressure at  
which the inlet should be set. The inlet should not be set at or below 3.8 psig.  
If you decide to use a typical configuration, note the restrictor dimensions  
from Table 2 and proceed to “Restrictor and Column Installation" on page 35.  
Splitting to an MSD  
Note that the maximum column flows for an MSD are quite low. This limit is  
imposed by the rating of the turbo or diffusion pump. Configurations with  
split ratios greater than 1 can be used but peak broadening and/or tailing  
should be expected. They are shown in the configuration tables more as a  
caution than as a recommendation.  
In practice, the column flow can be set to within 0.5 mL/min of the total flow  
if necessary. For example, the 1:1 split to an MSD with a performance turbo  
pump running a method programmed to 300 °C should have a column flow of  
no more than 2.3 mL/min ([1.4 + 1.4] – 0.5) at 300 °C.  
Split ratios to the MSD greater than 1 are very limited due to these flow  
considerations and should be avoided if possible.  
28  
Installation and Operation Guide  
                 
Splitter Configurations  
3
Custom Configurations  
The CD supplied with this kit contains three software tools:  
Effluent Splitter Calculator (with Makeup) Calculates dimensions (length and  
inside diameter) of restrictors to obtain a desired split ratio (Figure 11).  
Figure 11 Effluent Splitter Calculator  
Installation and Operation Guide  
29  
             
3
Splitter Configurations  
GC Method Translation Converts an analytical method from one set of  
operating conditions to another (Figure 12).  
Figure 12 GC Method Translation  
30  
Installation and Operation Guide  
   
Splitter Configurations  
3
Column Pressure/Flow Calculator Calculates flows and pressures for a given  
set of column (or restrictor) dimensions (Figure 13).  
Figure 13 Column Flow/Pressure Calculator  
These tools allow you to perform all the calculations needed to create a  
custom splitter configuration. We recommend that you load the CD software  
into your PC.  
1 Insert the CD into the drive and click the Start icon in the bottom left of the  
screen. Select Run and type X:\ Setup, where X is the letter assigned to the  
CD drive.  
2 Click Start, then select Programs and the program you wish to run.  
Installation and Operation Guide  
31  
     
3
Splitter Configurations  
Restrictor id and length  
1 Run the Effluent Splitter Calculator and enter the following information.  
The calculator provides a list of possible restrictors.  
Column flow. Use the ChemStation, GC, Flow Calculator, or Method  
Translation Software to determine the column flow in mL/min (with the  
column outlet at 3.8 psig) at the initial oven temperature.  
Initial oven temperature. This is the temperature setpoint for an  
isothermal method or the initial temperature for a programmed method.  
Carrier gas type. Enter Helium, Hydrogen, Nitrogen, or Argon.  
Detectors 1 and 2 operating pressure (psia). The operating pressure  
must be in absolute units. Most detectors (FID, TCD, ECD, NPD, and  
FPD) operate at atmospheric pressure (14.696 psia). Exceptions are the  
MSD (0 psia) and AED (16.196 psia).  
Flow Ratio of Detector 2 to Detector 1. This is the desired split ratio  
between the two detectors. Usually this number is 1, meaning the  
effluent divides equally between the detectors. This can be adjusted to  
higher values, but should normally not exceed five.  
Splitter (column outlet) pressure (psig). This is the desired pressure at  
which the splitter (and thus the end of the column) will operate. It can be  
set between 2 and 4 psig, but is usually set to 3.8 psig. This number can  
be varied to obtain an acceptable combination of restrictors that will  
have sufficient flow velocity to give good peak shapes.  
2 Choose the id tubing that gives a length closest to (and at least) 0.3 m for  
most detectors and 0.8 m for MSDs. The green fields with tubing diameters  
in mm can be edited if you have other sizes of deactivated tubing available.  
32  
Installation and Operation Guide  
   
Splitter Configurations  
3
Maximum and minimum flows  
The maximum suggested flow for MSDs depends on the vacuum pump used.  
For diffusion pump and standard turbo systems, the flow should not exceed  
2 mL/min. For performance turbo systems, the flow should not exceed  
4 mL/min. These flow limits restrict the column flows and split ratios that can  
be used with MSDs.  
Make sure that the flow through each restrictor tube is at least equal to the  
suggested minimum flow in Table 4. Restrictors that fail this test will still  
work, but peak broadening and/or tailing may result.  
Table 4  
Suggested minimum restrictor flows  
Minimum carrier gas flow, mL/min  
Restrictor internal  
diameter, mm  
Helium  
0.400  
0.720  
0.800  
1.000  
1.280  
1.800  
2.120  
Hydrogen  
0.500  
Nitrogen  
0.125  
0.225  
0.250  
0.313  
0.400  
0.563  
0.663  
Argon  
0.110  
0.198  
0.220  
0.275  
0.352  
0.495  
0.583  
0.10  
0.18  
0.20  
0.25  
0.32  
0.45  
0.53  
0.900  
1.000  
1.250  
1.600  
2.250  
2.650  
1 The makeup flow is listed in cell B 31 of the effluent splitter calculator. You  
should have at least 0.5 mL/min for stable pressure regulation. Note that  
this value will decrease as the oven temperature programs up.  
2 Use the Column Pressure/Flow Calculator to determine the flow through each  
restrictor at the maximum oven temperature of the method, add them and  
subtract the calculated column flow at that temperature. This value should  
be greater than 0.5 mL/min.  
Installation and Operation Guide  
33  
           
3
Splitter Configurations  
Column outlet pressure  
The 6890 GC needs to know the pressure at the end of the column to be able to  
calculate column flows. Use either the GC keyboard or the ChemStation to set  
the outlet pressure for the column to 3.8 psig. The ChemStation screen where  
the column outlet pressure is set is shown in Figure 14.  
Figure 14 Column outlet pressure screen  
Inlet pressure  
If this is a method used previously, you may want to reset the inlet pressure to  
give similar retention times with the new column outlet pressure. Do this by  
calculating the inlet pressure needed to keep the void (holdup) time the same  
as the previous method. For constant inlet pressure methods, this will also  
keep the elution order the same. The Method Translation Software tool or the  
Flow Calculator tool can be used to do this calculation.  
34  
Installation and Operation Guide  
               
Splitter Configurations  
3
Restrictor and Column Installation  
Restrictors and the column exit are connected to the splitter assembly using internal nuts  
and SilTite ferrules. See “Swaging SilTite Ferrules” on the CD for details.  
NOTE  
Install the column  
1 Hang the analytical column on the column clips. The clips hold the outside  
of the wire “basket” that supports the column. Adjust the clips if necessary.  
2 Connect the column to the inlet fitting.  
Connect the splitter  
1 Connect the restrictors to the connectors on the splitter (Figure 15).  
Finger-tighten until just snug, then tighten with a wrench an additional 15°  
(Figure 16). Install the back restrictor first.  
2 Connect the restrictors to the appropriate detectors.  
3 Connect the column exit to the splitter. Tighten as you did the restrictors.  
Arrange the tubing (restrictors and column) so that it does not touch the oven walls.  
This could create a cold spot.  
CAUTION  
Installation and Operation Guide  
35  
           
3
Splitter Configurations  
To makeup  
gas supply  
Restrictor 2  
to Detector 2  
Restrictor 1  
to Detector 1  
Column exit  
Figure 15 Restrictor and column connections  
Do not overtighten the fittings. The dashed line in Figure 16 (about 15° clockwise from  
finger-tight) is usually enough.  
CAUTION  
36  
Installation and Operation Guide  
 
Splitter Configurations  
3
15°  
Figure 16 Tightening the connections  
Disconnect tubing from the splitter  
Loosen and remove the internal nut from the splitter fitting. Usually the  
tubing and ferrule will fall out of the fitting.  
Occasionally the ferrule will stick in the fitting. If this happens, use a pointed  
object like a pen or a paper clip and insert it in the ferrule release hole in the  
side of the fitting (Figure 17). Press firmly. The ferrule will click when it  
breaks free.  
Figure 17 Releasing a ferrule  
Installation and Operation Guide  
37  
       
3
Splitter Configurations  
Protect the column and restrictors  
Column and restrictor tubes with swaged metal ferrules can be disconnected  
and reconnected several times. To protect the tubing end, use one of the  
brass-sealing caps from the kit. Tighten to finger-tight plus 15 degrees.  
Protect the splitter  
Seal the ports of the splitter assembly with plugs when the splitter is not  
connected. This keeps particulates and contamination out. To make a plug, cut  
about 2 inches of the stainless steel wire and swage it as you would a column.  
Use the metal ferrule that fits 0.25-mm id columns. After swaging, clip the  
wire to within 0.5 mm of the ferrule end with a small high-quality wire cutter.  
Leave the excess wire on the other end to serve as a handle when removing the  
plug.  
38  
Installation and Operation Guide  
 
Agilent G3180B Splitter Kit  
Installation and Operation Guide  
4
An Example 40  
Calculate column flow 43  
Calculate ECD restrictor flow 44  
Calculate MSD restrictor flow 45  
Changing Columns Without Venting the MSD 46  
Backflushing the Column 47  
This chapter contains a worked-through custom configuration, plus some  
special topics.  
39  
Agilent Technologies  
 
4
Operation  
An Example  
Assume we have a method that uses an HP-5MS column (30 m × 250 µm  
id × 0.25-µm film thickness) to measure pesticides with an MSD. The initial  
ο
oven temperature is 70 °C and is programmed to 280 C. The method is run in  
constant pressure mode at 19.44 psig inlet pressure and the carrier gas is  
helium. The initial column flow listed by the ChemStation is 2.1 mL/min.  
We want to create a new splitter method with the column effluent split 1:3  
between the ECD (detector 1) and an MSD (detector 2). We would also like to  
preserve the retention times and relative elution order in the new method.  
Column flow  
Since the column outlet pressure will be much higher in the new method, the  
first step is to calculate the new inlet pressure and the resulting column flow.  
The Method Translation software (Figure 18) is useful for this. Use the None  
mode and check the button to make the hold-up times the same.  
40  
Installation and Operation Guide  
         
Operation  
4
Figure 18 Calculating column flow  
The outlet pressure entered for the new splitter method must be in absolute  
pressure units. Since the outlet of the column will be 3.8 psig, we need to  
convert this to psia for the method translator. Absolute pressure = gauge  
pressure + 14.696. Hence, 3.8 + 14.696 = 18.496 will be entered.  
The calculated inlet pressure for the new splitter method is 30.93 psig and the  
new column flow is 3.09 mL/min.  
Installation and Operation Guide  
41  
         
4
Operation  
Select restrictors  
Start up the spreadsheet "splitter_calc.xls" in Excel. We will choose to have  
2 mL/min go to the ECD initially. With a split ratio of 3, this will send  
6 mL/min to the MSD. This flow is acceptable with a performance turbo  
system but will give somewhat degraded detection limits. Fill in the input  
column as shown (Figure 19) with the ECD assumed to be Detector 1 and the  
MSD as Detector 2.  
Figure 19 The Effluent Splitter calculator  
42  
Installation and Operation Guide  
     
Operation  
4
The calculator lists the lengths required for the different sizes of uncoated,  
deactivated, fused-silica, restrictor tubing available. Choose the id tubing that  
gives the shortest length of at least 0.3 m for most detectors and 0.8 m for  
MSDs. In this case 0.18-mm id is the choice, requiring 0.912 m for the ECD  
restrictor and 0.824 m for the MSD restrictor.  
Table 4 on page 33 shows that in both cases the flow is higher than the  
minimum 0.72 mL/min suggested for helium in 0.18-mm id tubing.  
Calculate column flow  
To find the makeup flow at 280 °C, first find the column flow at 280 °C. The  
Flow Calculator software (Figure 20) requires that the output pressure be  
entered in psia. Therefore 18.496 psia (3.8 psig) is entered.  
Figure 20 Column flow calculation  
The column flow drops to 1.38 mL/min at 280 °C.  
Installation and Operation Guide  
43  
     
4
Operation  
Calculate ECD restrictor flow  
The flow through the ECD restrictor at 280 °C is calculated to be 0.88 mL/min  
(Figure 21).  
Figure 21 ECD restrictor flow calculation  
This flow is higher than the minimum 0.72 mL/min suggested for helium in  
0.18-mm id tubing.  
44  
Installation and Operation Guide  
     
Operation  
4
Calculate MSD restrictor flow  
The flow through the MSD restrictor (Figure 22) at 280 °C is:  
Figure 22 MSD restrictor flow calculation  
The flow to the MSD at 280 °C is 2.65 mL/min. This flow is higher than the  
minimum 0.72 mL/min suggested for helium in 0.18-mm id tubing. The  
calculated makeup flow is then [0.88 + 2.65] – 1.38 = 2.15 mL/min. This should  
work well.  
The configuration can now be installed and used.  
Installation and Operation Guide  
45  
       
4
Operation  
Changing Columns Without Venting the MSD  
For systems that use an MSD attached to the splitter, one added advantage is  
the GC column can be changed without venting the MSD. When the column is  
disconnected from the splitter plate, the makeup gas purges air out of the  
fitting, preventing air from reaching the MSD.  
To change columns with the splitter, the recommended steps are:  
1 Cool down the inlet to which the column to be removed is connected.  
2 Disconnect the column from the splitter plate.  
3 Immediately install a plug in the plate where the column was connected.  
4 Change column in the inlet and turn on carrier gas to purge air from the  
column.  
5 Preswage metal ferrule on the outlet end of the column.  
6 Remove plug from the connector.  
7 Connect the new column to the splitter.  
46  
Installation and Operation Guide  
     
Operation  
4
Backflushing the Column  
One useful feature available with EPC control of the makeup is the ability to  
backflush unwanted higher boiling analytes from the column. Use of this  
feature requires that the split/splitless inlet be used. Backflushing reduces the  
hold at the end of the run to clean out the column.  
To backflush, the splitter makeup pressure is time-programmed to rise rapidly  
after elution of the last peak of interest while the inlet pressure decreases  
rapidly. These pressure changes reverse the flow through the column. Heavy  
materials are then carried out the split vent of the inlet.  
The inlet pressure is programmed to decrease to 0.5 psig. The makeup  
pressure is programmed to rise to a maximum pressure determined by the  
detectors and cleanout temperature used. Using the example from above, the  
MSD will limit the flow, and thus pressure, that can be used for backflushing.  
The flow allowed to go to the MSD (with a performance turbo) must be  
8 mL/min or less. The backflushing conditions must be calculated to not  
exceed this. We need to use the MSD restrictor tubing dimensions and the  
backflushing temperature to find the backflushing pressure.  
Installation and Operation Guide  
47  
       
4
Operation  
The restrictor to the MSD was 0.824 m of 0.18-mm id tubing. The backflushing  
temperature used here is the hold temperature at the end of the run in the  
original method (280 °C). The flow calculator (Figure 23) shows that the  
makeup pressure can be programmed to 17.4 psig at 280 °C.  
Figure 23 Column backflush flow calculation  
The time required for complete backflushing of heavy materials is then  
determined empirically. Blank runs after samples with different backflush  
hold times are used to determine the minimum time to remove all heavy  
material.  
48  
Installation and Operation Guide  
 
Index  
Excel, 12  
Peak broadening, 26, 28  
Plugs, 38  
Pressure controller, 12  
A
Absolute pressure, 41  
Aux EPC, 27  
Auxiliary Pressure controller, 24  
F
Ferrule  
release hole, 37  
Flow  
column, 10, 26, 27  
B
R
Backflushing, 47  
detector, 26  
makeup, 10  
Restrictor, 8, 27  
dimensions, 26  
flow, 44, 45  
C
Changing columns, 46  
Column  
Retention times, 34  
connections, 35  
flow, 10, 26, 27, 43  
outlet pressure, 34  
Column clips, 35  
Column effluent, 27  
Configuration, 9  
GC Method Translation, 30  
GC requirements, 12  
Software tools  
Column Pressure/Flow Calculator, 31  
GC Method Translation, 30  
loading, 31  
custom, 25, 29  
typical, 25, 26  
Inlet pressure, 28, 34, 47  
Split ratio, 9, 26, 28, 29  
Constant flow mode, 27  
Constant pressure mode, 28  
Constant pressure operation, 10  
Custom configuration, 29  
Example, 40  
Makeup flow, 33  
T
Makeup supply, 24, 27  
Metal ferrules, 10  
Method Translation, 34, 40  
Microfluidic plate, 10  
MSD, 10, 28, 33, 40, 45, 46  
Tailing, 28  
Turbo pump, 26, 28  
D
Detector  
Above atmospheric pressure, 9  
Atmospheric pressure, 9  
Below atmospheric pressure, 9  
flow rate, 26  
operating pressure, 26  
Diffusion pump, 26, 28  
Outlet pressure, 41  
P
E
Parameter calculation, 11  
Parts supplied, 13  
PCM, 12, 27  
Effluent Splitter Calculator, 29, 32, 42  
Electronic pneumatic control, 10  
Installation and Operation Guide  
49  
Installation and Operation Guide  
50  
Agilent Technologies  
© Agilent Technologies, Inc.  
Printed in USA, April 2006  
G3180-90120  

Nexstar NST 200S3 BK User Manual
Mr Coffee IDS40 User Manual
Motorola MOTORAZR 6809504A81 A User Manual
Motorola MOTO Manual Number_ User Manual
Memorex MD5585 User Manual
Lindy 42782 User Manual
Huawei HW M750 User Manual
Echo Bear Cat CH45390 User Manual
Black & Decker CM2070B User Manual
Audiovox 8910 User Manual