HP (Hewlett Packard) Data Explorer 4 Series User Manual

Data ExplorerSoftware  
Version 4 Series Software  
User Guide  
Table of Contents  
How to Use This Guide .............................................................. xi  
Chapter 1 Data Explorer™ Basics  
1.1  
1.2  
Overview .............................................................................. 1-2  
File Formats and Types ............................................................ 1-5  
1.2.1  
Software Applications Compatibility ....................................... 1-5  
1.2.2  
Data (.DAT) File Format ......................................................... 1-5  
1.3  
1.4  
Parts of the Data Explorer Window............................................. 1-11  
Customizing the Data Explorer Window....................................... 1-17  
1.4.1  
1.4.2  
1.4.3  
1.5  
1.6  
Setting Graphic Options .......................................................... 1-23  
1.5.1  
1.5.2  
Customizing Graphic Options ................................................1-24  
Reverting to Previous Graphic Options .................................1-29  
1.6.1  
(Mariner Data Only) ..............................................................1-30  
(Mariner Data Only) ..............................................................1-33  
1.6.3  
1.6.4  
1.6.5  
Converting to and Exporting ASCII Data ..............................1-34  
Importing a Trace in ASCII Format ........................................1-35  
Extracting and Saving Information  
from .DAT, .RSD, and .RCD Files ..........................................1-36  
1.6.6  
Copying from Data Files ........................................................1-38  
Data ExplorerSoftware Users Guide  
iii  
Table of Contents  
2.1  
2.1.1  
2.1.2  
2.1.3  
2.1.4  
2.1.5  
2.1.6  
Displaying Voyager Chromatograms ...................................... 2-7  
Viewing Read-Only Files ........................................................ 2-7  
Moving Between Open Files .................................................. 2-8  
Closing Data Files .................................................................2-10  
2.2  
2.3  
2.4  
Adjusting the Display Range..................................................... 2-11  
2.4.1  
2.4.2  
2.4.3  
2.4.4  
Zooming, Centering, and Customizing a Trace ......................2-14  
Duplicating a Trace ...............................................................2-15  
Dividing the Active Trace ......................................................2-15  
Adding Traces from  
2.4.5  
2.4.6  
2.4.7  
2.4.8  
2.4.9  
Overlaying Traces .................................................................2-24  
2.5  
2.5.1  
Copying Traces from Multiple Data Files to a Window ...........2-37  
2.6  
2.7  
Saving, Opening, and Deleting .DAT Results ................................ 2-38  
Exporting, Opening, and Deleting .RCD and .RSD Results Files  
(Mariner Data Only)................................................................ 2-39  
2.8  
Saving, Opening, and Deleting .SPC Results Files  
(Mariner Data Only)................................................................ 2-40  
iv  
Applied Biosystems  
Chapter 3 Peak Detection and Labeling  
3.1  
3.1.1  
3.1.2  
3.2  
3.2.1  
3.2.2  
3.2.3  
3.2.4  
3.2.5  
Strategy for Voyager Peak Detection ...................................... 3-8  
Setting Peak Detection Parameters ......................................3-11  
Peak Detection Parameter Descriptions ................................3-19  
Charge State  
3.3  
3.3.1  
3.3.2  
3.3.3  
3.3.4  
Displaying the Peak List ........................................................3-37  
Inserting Peaks in the Peak List ............................................3-39  
Saving the Peak List .............................................................3-40  
Sorting, Filtering, and Printing the Peak List .........................3-42  
3.4  
3.5  
3.5.1  
3.5.2  
3.5.3  
Charge State Labels .............................................................3-53  
Setting Chromatogram and Spectrum Peak Labels ...............3-54  
Setting Custom Peak Labels .................................................3-61  
3.6  
3.7  
Process that Occurs During Peak Detection,  
Centroiding, and Integration ..................................................... 3-67  
Default Peak Detection Settings ................................................ 3-71  
Data ExplorerSoftware Users Guide  
v
Table of Contents  
4.1  
4.2  
Creating an Extracted Ion Chromatogram ..................................... 4-5  
4.2.1  
4.2.2  
4.3  
4.4  
4.5  
4.6  
4.7  
Noise Filtering/Smoothing ........................................................ 4-17  
Adding and Subtracting Raw Spectra Within a Data File.................. 4-20  
Displaying MS Method Data (Mariner Data Only) ........................... 4-23  
4.7.1  
Using Baseline Offset ...........................................................4-27  
4.7.2  
Using Baseline Correction .....................................................4-29  
4.8  
Using UV Trace Offset (Mariner Data Only) .................................. 4-30  
5.1  
5.2  
5.3  
Overview .............................................................................. 5-2  
Creating a Combined Spectrum ................................................. 5-4  
5.3.1  
5.3.2  
5.3.3  
5.3.4  
5.3.5  
5.3.6  
Creating or Modifying a Calibration Reference File (.REF) ....5-17  
5.4  
Automatic Calibration.............................................................. 5-26  
5.4.1  
5.4.2  
5.4.3  
Overview of Automatic Calibration ........................................5-26  
Importing and Specifying Automatic Calibration Settings .......5-29  
Automatically Calibrating (Mariner Data Only) .......................5-34  
vi  
Applied Biosystems  
Table of Contents  
5.5  
5.6  
5.7  
5.8  
Centroiding .......................................................................... 5-36  
Mass Deconvolution (Mariner Data Only)..................................... 5-37  
Adjusting the Baseline ............................................................ 5-45  
5.8.1  
5.8.2  
Using Baseline Correction .....................................................5-47  
5.9  
Truncating a Spectrum ............................................................ 5-56  
or Different Data Files (Dual Spectral Trace Arithmetic)................... 5-64  
6.1  
Using the Elemental Composition Calculator ................................. 6-2  
6.1.1  
Determining Elemental Composition ...................................... 6-2  
6.1.2  
Setting Limits ......................................................................... 6-7  
6.2  
6.3  
6.4  
6.5  
6.6  
6.7  
Using the Isotope Calculator..................................................... 6-13  
Using the Signal-to-Noise Ratio Calculator................................... 6-23  
Using the Ion Fragmentation Calculator....................................... 6-25  
Using the Elemental Targeting Application ................................... 6-31  
6.7.1  
6.7.2  
6.7.3  
6.7.4  
6.7.5  
6.7.6  
6.7.7  
6.7.8  
Deleting a Macro ...................................................................6-41  
Advanced Macro Editing .......................................................6-42  
Importing or Exporting Macros in DATAEXPLORER.VB6 ......6-43  
Running Macros Automatically  
When Opening and Closing Files ..........................................6-45  
Data ExplorerSoftware Users Guide  
vii  
Table of Contents  
7.1  
Mariner Data Examples ............................................................ 7-2  
7.1.1  
Improving Signal-To-Noise Ratio ............................................ 7-2  
7.1.2  
Deconvoluting and Evaluating  
7.1.3  
7.2  
Voyager Data Examples .......................................................... 7-11  
7.2.1  
7.2.2  
7.2.3  
Detecting and Labeling Partially Resolved Peaks ..................7-11  
Processing Before Calibrating to Optimize Mass Accuracy ...7-14  
Detecting Peaks from Complex Digests ................................7-18  
8.1  
8.2  
8.3  
Displaying PSD Data ............................................................... 8-2  
Calibrating a PSD Spectrum ..................................................... 8-10  
8.3.1  
8.3.2  
Checking Peak Detection ......................................................8-11  
Calibrating ............................................................................8-12  
Applying to Other Data Files .................................................8-20  
8.3.4  
8.3.5  
Creating PSD Calibration Reference (.REF) Files .................8-21  
Changing the Precursor Mass ...............................................8-23  
9.1  
9.2  
9.3  
9.4  
9.5  
9.6  
General Troubleshooting .......................................................... 9-3  
Processing, Tools, and Applications Troubleshooting....................... 9-6  
Calibration Troubleshooting...................................................... 9-10  
Printing Troubleshooting .......................................................... 9-14  
Peak Detection and Labeling Troubleshooting............................... 9-15  
viii  
Applied Biosystems  
Table of Contents  
Appendix A Warranty Information........................................ A-1  
Appendix B Overview of Isotopes........................................ B-1  
Appendix C Data Explorer Toolbox  
(Visual Basic Macros) ......................................................... C-1  
C.1 Overview .............................................................................. C-2  
C.2  
Preparing Data Before Accessing Macros ..................................... C-3  
C.3 Accessing the Macros .............................................................. C-4  
C.4  
C.5  
C.6  
C.7  
Using the Ladder Sequencing Toolbox ......................................... C-5  
Using the Peptide Fragmentation Toolbox ..................................... C-9  
Using the Polymer Analysis Toolbox .......................................... C-15  
Using MS Fit/MS Tag Toolbox .................................................. C-18  
Index  
Data ExplorerSoftware Users Guide  
ix  
Table of Contents  
x
Applied Biosystems  
How to Use This Guide  
1
How to Use This Guide  
Purpose of  
this guide  
The Applied Biosystems Data Explorer Software Users  
Guide describes processing and analyzing data with the  
Data Explorer software. You can use the Data Explorer  
software to analyze data collected on:  
MarinerWorkstations with Version 3.0 and later  
software  
Voyager-DE BiospectrometryWorkstations with  
Version 5.0 and later software  
Audience  
This guide is intended for novice and experienced Mariner  
or Voyager workstation users who are analyzing  
biomolecules.  
Structure of  
this guide  
The Applied Biosystems Data Explorer Software Users  
Guide is organized into chapters and appendixes. Each  
chapter page is marked with a tab and a header to help  
you find information.  
The table below describes the material covered in each  
chapter and appendix.  
Content  
Chapter 1, Data ExplorerBasics  
Describes file formats, file management, the  
parts of the Data Explorer window, and how  
to customize the Data Explorer software.  
Chapter 2, Using Chromatogram  
and Spectrum Windows  
Describes window and trace handling. Also  
describes saving results.  
and Labeling  
Provides background information on peak  
detection, centroiding, and integration.  
Describes peak detection, peak labeling,  
and peak deisotoping.  
Chapter 4, Examining  
Chromatogram Data  
Describes processing and analyzing  
chromatographic data.  
Data ExplorerSoftware Users Guide  
xi  
   
How to Use This Guide  
1
Chapter/Appendix  
Content  
Chapter 5, Examining Spectrum  
Data  
Describes processing and analyzing mass  
spectral data.  
Chapter 6, Using Tools and  
Applications  
Describes how to generate results using  
several tools and applications: the Centroid  
calculator, Elemental Composition  
calculator, Isotope calculator, Mass  
Resolution calculator, Ion Fragmentation  
calculator and Signal-to-Noise calculator.  
Also describes using the Macro Recorder  
and the Elemental Targeting Application.  
Chapter 7, Data Explorer Examples  
Includes specific examples for Mariner data  
and Voyager data. Examples include how to  
improve the signal-to-noise ratio for  
reserpine, deconvolute unresolved peaks in  
cyctochrome c (Mariner data), and label  
partially resolved peaks (Voyager data).  
Chapter 8, Viewing Voyager PSD  
Data  
Describes how to view, label, and calibrate  
PSD data.  
Includes symptoms and possible causes of,  
and corrective actions for potential system  
problems.  
Appendix A, Warranty  
Provides warranty and service information.  
Appendix B, Overview of Isotopes  
Includes background information you need  
for understanding isotopes.  
Appendix C, Data Explorer Toolbox  
(Visual Basic Macros)  
Describes loading Visual Basic® macros,  
preparing data, and running the macros.  
xii  
Applied Biosystems  
How to Use This Guide  
1
Conventions  
This guide uses the following conventions to make text  
easier to understand.  
Bold indicates user action. For example:  
Type 0 and press Enter for the remaining  
fields.  
Italic text denotes new or important words, and is also  
used for emphasis. For example:  
Before analyzing, always prepare fresh matrix.  
Notes, Cautions,  
Warnings, and  
Hints  
A note provides important information to the operator. For  
example:  
NOTE: If you are prompted to insert the boot diskette  
into the drive, insert it, then press any key.  
A caution provides information to avoid damage to the  
system or loss of data. For example:  
CAUTION  
Do not touch the lamp. This may damage the lamp.  
A warning provides information essential to the safety of  
the operator. For example:  
WARNING  
CHEMICAL HAZARD. Wear appropriate personal  
protection and always observe safe laboratory practices  
when operating your system.  
A hint provides helpful suggestions not essential to the  
use of the system. For example:  
Hint: To avoid complicated file naming, use Save First  
to Pass or Save Best Only modes.  
Data ExplorerSoftware Users Guide  
xiii  
How to Use This Guide  
1
Related  
documentation  
The related documents shipped with your system include:  
Mariner Workstation Users GuideUse this  
document to learn detailed information about the  
Mariner Workstation.  
Voyager-DE Biospectrometry Workstation  
Users GuideUse this document to learn detailed  
information about the Voyager-DE Workstation.  
Printer documentation (depends on the printer you  
purchase)Use this documentation to set up and  
service your printer.  
Microsoft ® Windows NT ® Users Guide and  
related documentsUse this guide to learn detailed  
information about the Microsoft Windows NT user  
interface.  
Send us your  
comments  
We welcome your comments and suggestions for  
improving our manuals. You can send us your comments  
in two ways.  
On the web at:  
www.appliedbiosystems.com/about/contact.html  
By e-mail at:  
TechPubs@appliedbiosystems.com  
xiv  
Applied Biosystems  
   
Chapter  
1
1 Data Explorer™  
Basics  
This chapter contains the following sections:  
1.1 Overview ................................................... 1-2  
1.2 File Formats and Types ............................. 1-5  
1.3 Parts of the Data Explorer Window .......... 1-11  
1.4 Customizing the Data Explorer Window ... 1-17  
1.5 Setting Graphic Options .......................... 1-23  
1.6 Managing Files ........................................ 1-30  
Data ExplorerSoftware Users Guide  
1-1  
       
Chapter 1 Data ExplorerBasics  
1.1 Overview  
1
Description The Data ExplorerVersion 4.0 processing software is  
graphical software that you use to analyze, calibrate, and  
report data. You can use the Data Explorer software to  
analyze data collected on:  
MarinerWorkstations  
Voyager-DE BiospectrometryWorkstations  
NOTE: Application systems that automatically acquire  
and process data, such as Mariner High Throughput  
Analysis Option (CombiSolve) and Proteomics  
Solution 1Option, require specific versions of Data  
Explorer software.  
Features Data Explorer software includes a suite of tools and  
processing options to allow you to graphically and interactively  
manipulate chromatographic and mass spectral data. For  
example, you can:  
Smooth and noise-filter data.  
Automatically and manually calibrate spectral data.  
Set peak detection parameters and custom labels for  
regions of the trace. Detected peaks can be evaluated  
for charge-state determination according to  
user-defined parameters.  
Determine elemental composition, theoretical isotope  
distributions, resolution, signal-to-noise ratio, and  
fragment ions.  
Perform target compound analysis (elemental  
targeting).  
Customize windows, toolbars, and traces.  
Create scripts and macros to automate your work using  
the Macro Recorder and Visual Basic® Editor.  
1-2  
Applied Biosystems  
   
Overview  
Starting and To start the Data Explorer software from the Windows NT  
desktop, double-click the Data Explorer icon on the desktop.  
The Data Explorer window opens.  
exiting the  
software  
1
The Data Explorer window is blank with only a few menus  
displayed until you open a data file.  
Figure 1-1 shows the Data Explorer main window with a  
Mariner data file open. Figure 1-2 shows the Data Explorer  
main window with a Voyager data file open.  
To exit the software, select Exit from the File menu in the Data  
Explorer window. The Data Explorer software closes.  
Figure 1-1 Data Explorer Window with Mariner Data  
Data ExplorerSoftware Users Guide  
1-3  
   
Chapter 1 Data ExplorerBasics  
1
Figure 1-2 Data Explorer Window with Voyager Data  
Default colors The default colors are different for Mariner and Voyager:  
labels  
VoyagerWhite background, blue traces, and red  
labels  
You can customize the default colors as needed. See  
Section 1.5.1, Changing Background Color.  
NOTE: For consistency, all Mariner and Voyager screen  
examples in the following sections of this Users Guide are  
shown with a white background.  
1-4  
Applied Biosystems  
   
File Formats and Types  
1.2 File Formats and Types  
1
This section describes:  
Software applications compatibility  
Data (.DAT) file format  
You can use the Data Explorer Macro Recorder function to  
create Visual Basic scripts to automate tasks. You can also  
use the Visual Basic Editor directly to create more complex  
programs customized to suit your needs. For more  
information, see Section 6.7, Using the Macro Recorder.  
Additionally, you can convert data to ASCII format for import  
into other software applications or import ASCII results. For  
more information, see Section 1.6.3, Converting to and  
Exporting ASCII Data, and Section 1.6.4, Importing a Trace in  
ASCII Format.  
1.2.2 Data (.DAT) File Format  
.DAT file format Data generated by Mariner and Voyager systems is stored in  
.DAT file format. The .DAT file format incorporates all  
information about how a data file was acquired and processed  
into a single file. This format improves data processing and  
data storage efficiency.  
Data files can contain spectra from a single acquisition or from  
multiple acquisitions (for example, multiple spectral data from  
a Voyager acquisition or multiple injection results from a  
Mariner CombiSolv run).  
Data ExplorerSoftware Users Guide  
1-5  
             
Chapter 1 Data ExplorerBasics  
Mariner .SPC file In Mariner software versions earlier than version 3.0, data files  
are stored in .SPC format. You can view and process .SPC  
format  
1
files in Data Explorer, or you can convert these files to .DAT  
format. For information about the differences between the  
.DAT and the .SPC formats, see the MarinerWorkstation  
Users Guide.  
Voyager .MS file In Voyager software versions earlier than version 5.0, data  
files are stored in .MS, .MSF, .MSA, and .MSB format. You can  
view and process .MS, .MSF, .MSA, and .MSB files in Data  
Explorer. You cannot convert them to .DAT format.  
format  
Data Explorer software.  
Extracting You can also store parameter settings in separate files by  
extracting information from a .DAT file as needed for use with  
other files. For more information, see Section 1.6.5, Extracting  
and Saving Information from .DAT, .RSD, and .RCD Files.  
information from  
.DAT files  
The types of information stored in a .DAT file are described  
below.  
Table 1-1 Information Stored In a .DAT File  
Category  
File Type  
File Content  
Settings  
.BIC  
Instrument settings for controlling the operation of the  
mass analyzer. For more information, see the:  
Mariner Workstation Users Guide  
Voyager Biospectrometry Workstation Users  
Guide  
1-6  
Applied Biosystems  
           
File Formats and Types  
Table 1-1 Information Stored In a .DAT File (Continued)  
1
Category  
File Type  
File Content  
Settings  
.MSM  
(Mariner  
only)  
MS Method settings, if data was acquired using an  
.MSM file.  
(continued)  
NOTE: To access the instrument settings used to  
acquire each spectrum in an MS Method, you must first  
extract the .MSM file from the data file, then export the  
the MS Method editor. For more information on  
exporting a .BIC file from an .MSM file, see the Mariner  
.SET  
.LBC  
.LBS  
.CAL  
Graphic and processing settings. See Section 1.4.2,  
Customizing Processing and Graphic Settings (.SET).  
Display  
Process  
Chromatogram label information. See Section 3.5.3,  
Setting Custom Peak Labels.  
Custom Peak Labels.  
Calibration constants generated by mass calibration.  
For more information, see Exporting .BIC, .MSM, and  
to additional fileson page 5-16.  
.CTS  
Processed trace that you access by selecting  
Processing History from the Display menu. For  
Rearranging Traces (Processing History).  
NOTE: You select the name of the trace from the  
Processing History menu. You do not directly select a  
.CTS file. To purge or disable .CTS files, see Setting  
Processing History optionson page 2-23.  
Data ExplorerSoftware Users Guide  
1-7  
         
Chapter 1 Data ExplorerBasics  
Additional files Additional file types you may see on your system are  
1
described below.  
types  
Table 1-2 Additional File Types  
Category  
File Type  
File Content  
Data  
.PKT  
Text file containing a chromatogram or a spectrum peak  
list that you can save from the Output window. See  
Output windowon page 1-15.  
.TXT  
Data file exported to an ASCII text file. See  
Section 1.6.3, Converting to and Exporting ASCII Data.  
Data  
.SPC  
(Mariner  
only)  
Data file format for files acquired before Version 3.0 of  
the Mariner Instrument Control Panel.  
NOTE: Voyager data files in .SPC format have a file  
structure different from Mariner data files in .SPC  
format, and are not supported in the Data Explorer  
software.  
.MS  
(Voyager  
only)  
Data file format for files acquired before Version 5.0 of  
the Voyager Instrument Control Panel.  
.MSA and  
.MSF  
(Voyager  
only)  
PSD data file format for composite and fragment files  
acquired before Version 5.0 of the Voyager Instrument  
Control Panel.  
NOTE: The Data Explorer software cannot generate  
composite spectra from .MSF fragment files.  
.MSB  
(Voyager  
only)  
Baseline-corrected data file format for composite and  
fragment files acquired before Version 5.0 of the  
Voyager Instrument Control Panel.  
Procedure  
.TUN  
AutoTune method. See the Mariner Workstation Users  
(Mariner  
only)  
Guide.  
1-8  
Applied Biosystems  
 
File Formats and Types  
1
Category  
File Type  
File Content  
Reference  
.REF  
List of masses to select from during calibration. See  
Creating and saving a calibration reference fileon  
page 5-18.  
Process  
.RCT  
(Mariner  
only)  
Mariner .SPC format data file (versions earlier  
than 3.0) in the Data Explorer software after a  
chromatogram is manually processed.  
See Section 2.8, Saving, Opening, and Deleting  
.SPC Results Files (Mariner Data Only).  
Mariner .DAT format data file (version 3.0 and later)  
in the Mariner Instrument Control Panel for a  
snapshot of chromatogram data.  
NOTE: Results for a .DAT file are stored within the .DAT  
file, not as a separate file.  
.RST  
Mariner .SPC format data file (versions earlier  
than 3.0) in the Data Explorer software after a  
spectrum is manually processed.  
See Section 2.8, Saving, Opening, and Deleting  
.SPC Results Files (Mariner Data Only).  
Voyager .MS format data file in the Data Explorer  
software after a spectrum is manually processed.  
Mariner .DAT format data file (version 3.0 and later)  
in the Mariner Instrument Control Panel for a  
snapshot of spectrum data.  
NOTE: Results for a .DAT file are stored within the .DAT  
file, not as a separate file.  
Data ExplorerSoftware Users Guide  
1-9  
Chapter 1 Data ExplorerBasics  
Table 1-2 Additional File Types (Continued)  
1
Category  
File Type  
File Content  
Process  
.RCD  
Section 2.7, Exporting, Opening, and Deleting .RCD and  
.RSD Results Files (Mariner Data Only).  
(continued)  
.RSD  
Spectrum results file exported from .DAT files. See  
Section 2.7, Exporting, Opening, and Deleting .RCD and  
.RSD Results Files (Mariner Data Only).  
1-10  
Applied Biosystems  
   
Parts of the Data Explorer Window  
1.3 Parts of the  
Data Explorer Window  
1
This section describes:  
Overview  
Toolbar  
Chromatogram and Spectrum windows  
Tabs for data files  
Data file names  
Output window  
Overview Figure 1-3 shows the Data Explorer window with Mariner data.  
Toolbar  
Chromatogram  
window  
Data file name  
Spectrum  
window  
Tabs for  
open data files  
Output  
window  
Figure 1-3 Parts of the Data Explorer Window  
Data ExplorerSoftware Users Guide  
1-11  
       
Chapter 1 Data ExplorerBasics  
Toolbar The toolbar contains buttons that access Data Explorer  
functions.  
1
For a description of a toolbar button, place the cursor on the  
button. A brief description of the button (ToolTip) is displayed  
below the button.  
For information on adding or removing toolbar buttons, see  
Customizing toolbarson page 1-21.  
Chromatogram Refer to the following tables for descriptions of the types of  
data that you can display in the Chromatogram (CHRO) and  
Spectrum (SPEC) windows:  
and Spectrum  
windows  
Mariner dataTable 1-3  
Voyager dataTable 1-4  
Table 1-3 Mariner Data Displayed in Chromatogram and Spectrum Windows  
Window  
CHRO  
Mariner Data  
Displays:  
Total Ion ChromatogramIncludes the entire mass range saved in  
the data file.  
Extracted Ion Chromatogram (XIC) (optional)Includes only the  
signal response from a mass window or range  
Constant Neutral Loss Chromatogram (CNL) (optional)Extracts  
only the response from peaks that are separated by a selected mass  
difference.  
Optionally displays the following from Diode Array data (DAD):  
Total Absorbance Chromatogram (TAC)  
Channel (Ch)  
Extracted Absorbance Chromatogram (XAC)  
Can be displayed as % Intensity versus Retention Time or Spectrum  
number by selecting Traces from the Display menu, then selecting  
X Axis In, then selecting Spectrum Number or Time.  
1-12  
Applied Biosystems  
                       
Parts of the Data Explorer Window  
Table 1-3 Mariner Data Displayed in Chromatogram and Spectrum Windows  
(Continued)  
1
Window  
SPEC  
Mariner Data  
Displays the spectrum for the selected time in the TIC or TAC trace. By  
default, displays spectrum #1. The trace label includes DADfor spectra  
selected from TAC.  
Indicates Base Peak (BP) mass and intensity for the tallest peak in the  
spectrum. Displayed as % Intensity versus Mass-to-Charge (m/z). The  
right axis is scaled to the intensity of the base peak.  
Table 1-4 Voyager Data Displayed in Chromatogram and Spectrum Windows  
Window  
CHRO  
Voyager Data  
Window is not displayed by default.  
Optionally displays Total Ion Current (TIC) for multiple spectra .DAT files if  
you select Restore Chromatogram from the View menu.  
NOTE: DAD functions are not supported for Voyager data.  
SPEC  
Depends on the type of data file you open:  
Single spectrum filesDisplays the spectrum and labels the plot as  
spectrum #1.  
Multiple spectrum filesDisplays the spectrum for the selected  
time in the TIC trace (if displayed). By default, displays spectrum #1.  
PSD filesDisplays the composite spectrum and labels it as  
Stitched PSD. You can display segment spectra by clicking  
and  
.
Indicates Base Peak (BP) mass and intensity for the tallest peak in the  
spectrum. Displayed as % Intensity versus Mass-to-Charge (m/z). The  
right axis is scaled to the intensity of the base peak.  
Data ExplorerSoftware Users Guide  
1-13  
               
Chapter 1 Data ExplorerBasics  
Labels in the chromatogram or spectrum title identify the  
type of data displayed in the window. For a description of  
labels, see Section 2.4.10, Viewing Trace Labels.  
1
For more information on Chromatogram and Spectrum  
windows, see Chapter 2, Using Chromatogram and Spectrum  
Windows.  
Context-sensitive The commands displayed on the menus in the Data Explorer  
window depend on the window that is active when you select  
the menu:  
menus  
Only Mariner-related commands are displayed on  
menus when you open a Mariner .DAT file.  
Only Voyager-related commands are displayed on  
menus when you open a Voyager .DAT file.  
Only spectrum-related commands are displayed if you  
select menus when the Spectrum window is active.  
Only chromatogram-related commands are displayed if  
you select menus when the Chromatogram window is  
active. For Voyager multiple spectra .DAT files, UV  
Tabs for data files Chromatogram and Spectrum windows display a tab at the  
bottom (see Figure 1-3 on page 1-11) that allow you to switch  
between data files, or between Chromatogram and Spectrum  
windows for a Mariner data file. See Moving Between Open  
Fileson page 2-8.  
Data file names Data file names are displayed in the:  
Title bar of each window  
Tab for each window  
NOTE: If the full name of the data file does not fit in the tab,  
part of the name is displayed followed by .... To display  
the full name of the data file, place the cursor on the tab.  
The full name of the data file is displayed below the tab.  
1-14  
Applied Biosystems  
     
Parts of the Data Explorer Window  
Output window The Output window (see Figure 1-3 on page 1-11) displays  
tabs at the bottom that you can click to switch between the  
Output window  
tabs  
ResultDisplays results generated using commands on  
Only)  
Section 6.3, Using the  
Mass Resolution Calculator  
Section 6.4, Using the Signal-to-Noise Ratio  
Calculator  
Chro Peak List Displays results of chromatogram  
peak detection and integration. For more information,  
see Section 3.3, Peak List.  
Spectrum Peak ListDisplays results of spectrum  
peak detection, integration, and centroiding. For more  
information, see Section 3.3, Peak List.  
Sample InfoFor data files, displays the:  
Software version used for acquisition  
Acquisition time and sample comments entered  
when data was acquired  
For result files (.RCT, .RST, .RCD, .RSD), displays any  
processing functions that have been performed and  
saved in the result file.  
Data ExplorerSoftware Users Guide  
1-15  
                       
Chapter 1 Data ExplorerBasics  
Instrument SettingDisplays a list of instrument  
settings used to obtain the data. The settings are taken  
from instrument settings pages in the Instrument  
Control Panel. Also displays segments, event numbers,  
Mariner.  
1
Elemental AnalysisDisplays results for the  
Section 6.1, Using the Elemental  
Composition Calculator.  
Elemental TargetingDisplays results for the  
Elemental Targeting application. For information, see  
Section 6.6, Using the Elemental Targeting Application.  
Displaying, The Output window is automatically displayed when you  
generate results (for example, when you calculate resolution).  
clearing, and  
To display the Output window manually, select Output  
Window from the View menu.  
closing  
To clear the Output window, right-click the Output window,  
then select Clear Window.  
To close the Output window, deselect Output Window from  
the View menu, or right-click the Output window, then select  
Hide.  
Hint: To maximize the Chromatogram and Spectrum  
windows after hiding the Output window, click  
in the toolbar.  
or  
1-16  
Applied Biosystems  
       
Customizing the Data Explorer Window  
1.4 Customizing the  
1
Data Explorer Window  
This section includes:  
Setting default values  
Customizing Graphic and Processing settings  
Customizing toolbars  
1.4.1 Setting Default Values  
You can set defaults for most dialog boxes in the Data  
Explorer software by setting a value or selecting a button,  
closing the dialog box, then closing the data file you are  
viewing.  
The next time you open the dialog box, the last settings  
specified are displayed.  
1.4.2 Customizing Processing  
and Graphic Settings (.SET)  
This section includes:  
Overview of processing and graphic settings  
What settings contain  
Default processing and graphic settings  
Default graphic settings  
Customizing settings saved in a data file  
Making a copy of default .SET files before customizing  
Opening, customizing, and saving .SET files  
Applying a .SET file  
Data ExplorerSoftware Users Guide  
1-17  
                   
Chapter 1 Data ExplorerBasics  
Overview of Processing and graphic settings control how data is processed  
and displayed in the Data Explorer software. The last settings  
used are automatically saved in the data file when you close it.  
The next time you open the data file, you can select any of the  
following to apply:  
processing and  
graphic settings  
1
Settings from the data fileDescribed in Section 2.1,  
Opening and Closing Data Files.  
Default settingsDescribed below.  
Settings from a selected set fileYou can also apply  
processing and graphic settings that have been  
exported as stand-alone .SET files from other data files.  
See Applying a .SET fileon page 1-20.  
What settings Graphic settings include the attributes you set in Graphic  
Options, described in Section 1.5, Setting Graphic Options.  
contain  
Processing settings include:  
Peak detection parameters, described in Section 3.2.4,  
Peak Detection Parameter Descriptions  
Smoothing points, described in Section 5.7, Noise  
Filtering/Smoothing  
Automatic calibration settings, described in Section 5.4,  
Automatic Calibration  
NOTE: You can open a .SET file in Microsoft Notepad to  
view the complete file contents.  
Default The following default settings files (stand-alone .SET files) that  
contain both processing and graphic settings are provided on  
your system in the C:\MARINER\PROGRAM\Set Files  
directory or the C:\VOYAGER\PROGRAM\Set Files directory:  
processing and  
graphic settings  
VOYAGERLINEAR.SET  
VOYAGERREFLECTOR.SET  
VOYAGERPSD.SET  
For more information on Peak Detection settings, see  
Section 3.7, Default Peak Detection Settings.  
1-18  
Applied Biosystems  
               
Customizing the Data Explorer Window  
Additional .SET files that have been developed for detection of  
different types of data are included in the  
C:\VOYAGER\PROGRAM\SET FILES directory. The names of  
the .SET files indicate the type of data the files can be used  
for.  
1
The appropriate default settings for the type of data you open  
are automatically applied to a data file the first time you open it  
in Data Explorer. You can also manually apply these settings if  
desired. For information, see Applying a .SET fileon  
page 1-20.  
Default graphic Two additional default settings files (stand-alone .SET files)  
that contain graphic settings only are also provided on your  
system in the C:\MARINER\PROGRAM directory or the  
C:\VOYAGER\PROGRAM directory:  
settings  
DEFAULTWHITE.SET  
These .SET files contain the default graphic settings  
applied when you select White Background or Dark  
Background (after selecting Default from the Display  
menu). For information, see Section 1.5.1, Changing  
Background Color.  
Customizing You can customize settings saved in a data file by adjusting  
graphic or processing settings in the Data Explorer window.  
settings saved in  
a data file  
Settings are saved with a data file when you close the data  
file, and are automatically applied the next time you open the  
data file, if specified. For more information, see Section 2.1,  
Opening and Closing Data Files.  
You can also save the settings in a .SET file for use with other  
data files, as described in Saving .SET fileson page 1-37.  
Making a copy of All .SET files are user editable. However, before you edit the  
default processing/graphic .SET files (see page 1-18), make a  
copy of the original .SET file, in case you need to reload the  
settings for default peak detection. Copy the file using  
Windows NT® Explorer.  
default .SET files  
before  
customizing  
Data ExplorerSoftware Users Guide  
1-19  
         
Chapter 1 Data ExplorerBasics  
Opening, To open, customize, and save .SET files:  
customizing, and  
saving .SET files  
1
1. If you are customizing a default .SET file, make a copy  
of the original file before opening it. See Making a copy  
of default .SET files before customizingon page 1-19.  
2. Select Settings from the File Menu.  
3. Select one of the following:  
Restore Processing Settings  
Restore Graphic Settings  
Restore Graphic/Processing Settings  
4. In the Restore dialog box, select or type the name of the  
.SET file, then click OK.  
5. Customize processing and graphic settings as needed.  
For information on the contents of processing and graphic  
settings, see What settings containon page 1-18.  
6. To save settings, select Settings from the File menu,  
then select one of the following:  
Save Processing Settings As  
Save Graphic Settings As  
Save Graphic/Processing Settings As  
7. In the Save As dialog box, type the name of the .SET file,  
then click OK.  
Applying a .SET You can apply a .SET file two ways:  
file  
Select Recent Processing Settings from the File  
menu, then select a .SET file from the list of most  
recently used files  
NOTE: When you apply a .SET file from the Recent  
Processing Settings menu, only the process settings  
are applied.  
Select an option from the Settings command on the  
File menu  
1-20  
Applied Biosystems  
               
Customizing the Data Explorer Window  
To use the Settings option:  
1. Select Settings from the File menu.  
1
2. Select one of the following:  
Restore Processing Settings  
Restore Graphic Settings  
Restore Graphic/Processing Settings  
Revert to the Last Saved Graphic/Processing  
Settings  
3. If you select a Restore Settings option, select or type the  
name of the .SET file in the Restore dialog box, then  
click OK.  
Hint: You can also restore default processing and graphic  
settings when you open a file or files. See Section 2.1,  
Opening and Closing Data Files.  
1.4.3 Customizing Toolbars  
Customizing To customize the toolbar:  
toolbars  
1. Select Customize Toolbar from the Tools menu to  
display the Customize dialog box.  
2. To display or hide a toolbar section, click the Toolbars  
tab, then select or deselect a toolbar.  
3. To add a button to a toolbar, click the Commands tab,  
select the appropriate category, then click-drag the button  
to any toolbar in the main toolbar.  
Hint: To display a button description, click the button  
within the Customize dialog box. You can add buttons  
from any category to any toolbar. For example, you can  
add buttons from the File category to the Graph toolbar.  
Data ExplorerSoftware Users Guide  
1-21  
           
Chapter 1 Data ExplorerBasics  
4. To remove a button from a toolbar, click-drag the button  
from the toolbar.  
1
NOTE: The Customize dialog box must be displayed to  
click-drag a button from a toolbar.  
5. Click OK to close the Customize dialog box.  
Undocking The toolbar at the top of the Data Explorer window is divided  
into sections. A section is preceded by a double vertical bar.  
toolbars  
You can move (undock) each section of the toolbar within the  
Data Explorer window by click-dragging the double bar at the  
left side of the toolbar section.  
To move the toolbar section back to the top of the window,  
click-drag the toolbar back to the original position.  
You can display or hide each section individually, add or  
remove buttons on the toolbar, and rearrange the order of  
buttons displayed. To do so, you must have the Customize  
dialog box open, as described in the previous section.  
Creating You can create new toolbars by click-dragging buttons to a  
window area where there is no toolbar. You can then add  
buttons to the new toolbar as described in Customizing  
toolbars  
toolbarson page 1-21.  
1-22  
Applied Biosystems  
   
Setting Graphic Options  
1.5 Setting Graphic Options  
1
This section includes:  
Changing background color  
Customizing options  
Reverting to previous graphic options  
NOTE: Changes you make to Graphic Options are saved  
with the data file.  
1.5.1 Changing Background Color  
White or dark You can switch background color by selecting Default from  
the Display menu, then selecting:  
background  
White BackgroundDisplays blue traces and red  
labels by default. Default settings are contained in  
DEFAULTWHITE.SET.  
Dark BackgroundDisplays yellow traces and green  
labels by default. Default settings are contained in  
DEFAULTBLACK.SET.  
do not contain processing settings.  
You can customize the graphic settings associated with  
default settings if desired. For information, see Section 1.4.2,  
Customizing Processing and Graphic Settings (.SET).  
Data ExplorerSoftware Users Guide  
1-23  
             
Chapter 1 Data ExplorerBasics  
1.5.2 Customizing Graphic Options  
1
This section includes:  
Accessing graphic options  
Setting colors  
Setting line widths  
Setting data cursors  
Setting traces in line or bar mode  
Setting graphic compression  
Accessing To access graphic options:  
graphic options  
1. Display the trace of interest.  
2. From the Display menu, select Graphic Options.  
3. To use the graphic options settings for all traces, click  
Use same settings for all traces in the View Setup tab.  
4. Click a Graph Setup tab in the Graph and Plot Options  
dialog box (Figure 1-4).  
5. Set colors, line widths, data cursors, and graphic  
compression as described in the following sections.  
6. Click OK.  
1-24  
Applied Biosystems  
       
Setting Graphic Options  
1
Line or vertical  
bar traces  
Peak  
bounds  
Line width  
Data cursor  
Graphic  
compression  
Figure 1-4 Graphic Options Dialog Box—  
Graph Setup Tab  
Setting colors You can set colors manually or automatically.  
Manually To manually select the color of graph features (axis, peak  
bounds, tick labels, data cursor) and plot features (traces,  
peak labels):  
1. Select Graphic Options from the Display menu.  
2. Set colors in the Graph Setup tab (see Figure 1-4).  
Data ExplorerSoftware Users Guide  
1-25  
             
Chapter 1 Data ExplorerBasics  
When you manually set colors, note:  
1
Selections set to white (or line widths set to 0) may not  
print on certain printers.  
If you select different trace colors for multiple traces, only  
the color for the active trace is saved when you close the  
data file.  
Automatically Automatically assigning trace colors is useful when overlaying  
traces. To automatically assign trace colors:  
using Auto Color  
1. Select Graphic Options from the Display menu.  
2. Select Auto Color in the View Setup tab. The software  
assigns and displays trace colors when the traces are  
overlaid.  
When you use Auto Color:  
The active trace color stays at its original setting.  
Other trace colors are set based on the active trace color.  
For example, if the active trace is yellow, other traces are  
assigned the colors pale blue, pale green, and medium  
gray, which are the colors listed after yellow in the Trace  
color list, excluding white.  
NOTE: White is not used in Auto Color because white may  
not print on certain printers.  
Setting line You can control trace appearance by setting line widths. To set  
line widths:  
widths  
1. Select Graphic Options from the Display menu.  
2. Set line width in the Plot Setup section of the Graph Setup  
tab (see Figure 1-4 on page 1-25).  
NOTE: Line Widths of 0 or 1 (or lines set to the color white)  
may not print on certain printers. If traces do not print,  
change the line width (or color).  
1-26  
Applied Biosystems  
       
Setting Graphic Options  
Setting data To enable data cursors and set cursor labels and attributes:  
cursors  
1. Select Graphic Options from the Display menu.  
1
2. In the Data Cursor section of the Graph Setup tab (see  
Figure 1-4 on page 1-25), select Show Data Cursors,  
then select one of the following cursors from the Type  
drop-down list:  
XSingle vertical cursor  
YSingle horizontal cursor  
X-YVertical and horizontal cursors  
X-XTwo vertical cursors  
X-Y-XTwo vertical cursors and one horizontal cursor  
NOTE: If data cursors are displayed, they are printed when  
you print traces. To suppress data cursor printing, deselect  
Show Data Cursors before printing.  
To set the cursor mode, select the appropriate label type as  
described below:  
Label  
Type  
Options  
Y Label  
AbsoluteDisplays the number of counts.  
BP RelativeDisplays the % Intensity value relative to the base  
peak in the trace. Includes BP%marker on the cursor label.  
Display RelativeDisplays the % Intensity value relative to the  
largest peak in the current display range.  
X-X  
AbsoluteDisplays the Spectrum Number or Retention Time  
(chromatogram) or the Mass (spectrum).  
Label  
X-RelativeAvailable if you have two X cursors displayed. For the  
first cursor, displays the Spectrum Number or Retention Time  
(chromatogram) or the Mass (spectrum). For the second cursor,  
displays the appropriate value relative to the first cursor.  
For example, if you place the first X cursor in a spectrum at 100 m/z  
and the second X cursor at 80 m/z, the cursors are labeled  
100 and 20.  
Data ExplorerSoftware Users Guide  
1-27  
                 
Chapter 1 Data ExplorerBasics  
Setting traces in You can change the trace display from Line to Vertical Bars.  
Each vertical bar represents one data point.  
Line or Vertical  
1
Bar mode  
Vertical bar mode is useful when setting peak detection  
parameters to determine the number of points across a peak.  
Setting graphic NOTE: Graphic Compression mode is not saved as part of  
graphic settings. When you close a data file, it is  
automatically reset to the default Local Max setting.  
compression  
By default, all data is compressed when it is displayed on your  
computer screen. The degree of compression is determined  
by the number of data points in the data file and the resolution  
setting of your computer monitor.  
For example, assume that a data file contains 10,000 data  
points and needs compression to 1,000 data points to fit on  
your computer screen. Every 10 data points will be  
compressed into a single data point.  
To set graphic compression settings:  
1. Select Graphic Options from the Display menu.  
2. In the Graph Setup tab of the Graph and Plot Options  
dialog box (see Figure 1-4 on page 1-25), select a  
Graphic Compression mode:  
Local Max (default)Uses the maximum data  
point within the range of data points being  
compressed (range of 10 in the example above).  
AverageUses the average of the data points  
within the range of data points being compressed  
(range of 10 in the example above).  
Sum (Binning)Uses the sum of the data points  
within the range of data points being compressed  
(range of 10 in the example above).  
NOTE: Changing the graphic compression mode may  
alter the displayed intensities of peaks in a trace.  
1-28  
Applied Biosystems  
           
Setting Graphic Options  
1.5.3 Reverting to Previous Graphic Options  
1
You have two options to revert to previously used graphic  
options:  
Revert to Last Saved Graphic SettingReverts to  
the last graphic settings saved in the data file. Does not  
affect processing settings.  
To access, select Default from the Display menu, then  
select Revert to Last Saved Graphic Setting.  
Revert to Last Saved Graphic/Processing  
SettingsReverts to the last graphic and processing  
settings saved in the data file.  
To access, select Settings from the File menu, then  
select Revert to Last Saved Graphic/Processing  
Settings.  
Hint: Instead of applying the settings saved with a data file,  
you can apply the default settings stored in the default .SET  
file for your system (see page 1-18). To apply the default  
settings, close the data file, open the data file again, then  
select Use Default Settings in the Open dialog box. For  
more information, see Section 2.1, Opening and Closing  
Data Files.  
Data ExplorerSoftware Users Guide  
1-29  
 
Chapter 1 Data ExplorerBasics  
1.6 Managing Files  
1
This section describes:  
Converting .SPC file format to .DAT file format  
(Mariner only)  
Converting data from profile to centroid (Mariner only)  
Converting to and exporting ASCII data  
Importing a trace in ASCII format  
Extracting and saving information from .DAT, .RSD, and  
.RCD files  
Copying from data files  
1.6.1 Converting .SPC File Format to .DAT  
File Format (Mariner Data Only)  
This section includes:  
When to convert  
Before you begin  
Converting  
Viewing file properties  
Searching file properties  
NOTE: You cannot convert Voyager .MS files to .DAT  
format.  
When to convert You are not required to convert Mariner .SPC file format (for  
data acquired in software versions earlier than 3.0) to .DAT file  
format. However, the .DAT format allows you to store all  
information associated with the file (such as data, results,  
settings) in one file, simplifying file management.  
1-30  
Applied Biosystems  
           
Managing Files  
Before you begin Confirm that the .SPC and .CGM files are located in the  
same directory. Use Windows NT® Explorer to display the  
directory contents and to move the .SPC and .CGM files as  
necessary. If the .SPC and .CGM files are not in the same  
directory, when you open the .SPC file, a Failed to open  
chromatogram datamessage is displayed.  
1
To check that the .SPC and .CGM files are in the same  
directory:  
1. Select Open from the File menu.  
The Open dialog box is displayed.  
2. From the Files of type drop-down list, select All Files (*.*).  
A list of all files contained in the directory is displayed.  
3. Check that the .SPC and .CGM files are present.  
Converting In the Data Explorer window:  
1. Open or click the .SPC file to convert.  
2. Select Convert from the File menu.  
3. Select New Data Format.  
The Convert to .DAT Format dialog box is displayed  
(Figure 1-5).  
Figure 1-5 Convert to .DAT Format Dialog Box  
4. To add file property information (for example, Title, Author,  
or Keywords) to the file, click Set Property, enter the  
appropriate information, then click OK. For more  
information, see Viewing file propertieson page 1-32.  
Data ExplorerSoftware Users Guide  
1-31  
           
Chapter 1 Data ExplorerBasics  
The Convert to .DAT Format dialog box reappears.  
5. Click OK.  
1
A message box is displayed, showing the file name of  
the newly created .DAT file.  
6. Close the .SPC file and open the .DAT file before  
processing. If desired, manually delete the .SPC file and  
associated files.  
Viewing file File properties are accessible in Windows NT Explorer and  
provide general information about a data file without opening  
the file in the Data Explorer software.  
properties  
To view file properties:  
1. In Data Explorer, close the file of interest.  
2. In Windows NT Explorer, select the file, then right-click.  
3. Select Properties from the menu.  
4. Click the Summary tab.  
NOTE: If the Summary tab is not available, the file may  
be open in Data Explorer. Close the file and repeat the  
steps above.  
Searching file To search for a file based on file properties:  
properties  
1. In Windows NT Explorer, select Find from the Tools  
menu, then select Files or Folders.  
The Find: All Files dialog box is displayed.  
2. In the Name & Location tab, type or select a directory in  
the Look in text box.  
3. Select the Advanced tab.  
4. Select Data Explore Document from the Of type  
drop-down list.  
5. Type the information you are searching for (for example, a  
title or keyword) in the Containing text box, then click Find  
Now.  
1-32  
Applied Biosystems  
       
Managing Files  
1.6.2 Converting Data from  
1
Profile to Centroid (Mariner Data Only)  
Overview You can convert an entire data file from profile to centroid  
format. Centroid format files are smaller than profile format  
files.  
NOTE: Profile data is not automatically deleted when you  
convert to centroid data. You can delete the profile data file  
using Windows NT Explorer.  
Before you begin If the file to convert is in .SPC format, confirm that the .SPC  
and .CGM files are located in the same directory. For  
information see Before you beginon page 1-31. If the  
.SPC and .CGM files are not in the same directory, when  
you open the.SPC file, a Failed to open chromatogram  
datamessage is displayed.  
Converting to To convert from profile to centroid format:  
centroid  
1. Open or activate the .DAT or .SPC file to convert.  
2. Select Convert from the File menu, then select Centroid.  
The Save As dialog box is displayed. The software  
appends a -CTsuffix to the file name. This is the  
default file name. You can enter a new file name for the  
centroid file.  
3. Click Save.  
NOTE: The m/z range in a data file that is converted  
from profile to centroid is determined by the peak  
detection range set in Data Explorer, not the m/z range  
in the original data file. For more information, see  
Section 3.2, Peak Detection.  
Data ExplorerSoftware Users Guide  
1-33  
             
Chapter 1 Data ExplorerBasics  
1.6.3 Converting to and  
Exporting ASCII Data  
1
This section describes:  
Converting a data file to ASCII format  
Exporting a trace to ASCII format  
Converting a data You can convert a data file for use in a spreadsheet or another  
application and export and import single traces in ASCII  
format.  
file to ASCII  
format  
To convert an entire active data file to an ASCII text file:  
1. Open the data file to convert.  
2. From the File menu, select Convert, then select ASCII  
text.  
A Save As dialog box is displayed.  
3. Specify the name and destination for the file to be  
exported. By default, the software assigns a .TXT  
extension to the file.  
4. Click OK.  
Exporting a trace You can export a selected trace to ASCII format for display in  
Data Explorer software or for use in another application.  
to ASCII format  
To export trace data to ASCII format:  
1. With a data file open, select the trace window.  
2. From the File menu, select Export, then select  
Chromatogram or Spectrum (the menu item is context  
sensitive).  
A Save As dialog box is displayed.  
3. Specify the name and destination of the file to be  
exported. By default, the software assigns a .TXT  
extension to the file.  
4. Click OK.  
1-34  
Applied Biosystems  
             
Managing Files  
1.6.4 Importing a Trace in ASCII Format  
1
You can import trace data in ASCII format. If the file you are  
importing was originally exported using the Data Explorer  
software, you can import a spectrum trace only into the  
Spectrum window and a chromatogram trace only into the  
Chromatogram window. However, if the data is from another  
source, the software does not know the data type and you  
must make sure to import the data into the correct window  
type.  
Importing a Trace To import an ASCII trace:  
in ASCII Format  
1. Open a data file and activate a Chromatogram or  
Spectrum window.  
NOTE: You must open a data file before you can import  
ASCII data, even if the data file is unrelated to the  
imported data.  
2. Select Add/Remove Traces from the File menu, then set  
the Replace Mode by selecting one of the following:  
Replace the Active Trace  
Add a New Trace  
3. From the File menu, select Import, then select ASCII  
Chromatogram or ASCII Spectrum (the menu item is  
context sensitive to the selected window).  
An Open dialog box appears.  
4. Select the ASCII file to import.  
5. Click OK.  
The imported trace is displayed in the active window.  
Data ExplorerSoftware Users Guide  
1-35  
     
Chapter 1 Data ExplorerBasics  
CAUTION  
1
An imported ASCII format trace contains only the data  
points for the trace. The Sample Info and Instrument  
settings tabs in the Output window display data from the  
data file you opened in step 1. These tabs do not include  
information about the imported trace.  
1.6.5 Extracting and Saving Information  
from .DAT, .RSD, and .RCD Files  
Overview You can extract the following information from a .DAT data file,  
.RSD spectrum results file, and .RCD chromatogram results  
file, then save the information as a stand-alone file for use with  
other files:  
Instrument settings (.BIC)  
MS Method (.MSM) (Mariner data only)  
Calibration constants (.CAL)  
Processing/graphic settings (.SET)  
Spectrum or chromatogram peak labels (.LBS or .LBC)  
ASCII Spectrum and ASCII Chromatogram (.TXT)  
Exporting .BIC, To export Instrument settings (.BIC), MS Method settings  
(.MSM), or calibration constants (.CAL) from a .DAT or results  
file:  
.MSM, and .CAL  
files  
1. Open or activate the data (.DAT) or results (.RSD or  
.RCD) file.  
2. From the File menu, select Export, then select:  
CalibrationTo export the last applied  
calibration constants in a .CAL file.  
1-36  
Applied Biosystems  
                         
Managing Files  
NOTE: To export calibration constants used to  
1
acquire the data, select Mass Calibration from  
the Process menu, then select the Revert to  
Instrument Calibration before exporting. For  
more information, see Section 5.3.4, Reverting to  
Instrument Calibration.  
ConfigurationTo export .BIC or .MSM files  
(.MSMMariner data only).  
NOTE: To access the instrument settings for each  
spectrum in a data file acquired using a Mariner  
MS Method, you must first extract the .MSM file  
from the data file as described above. Then export  
the .BIC files from the .MSM file using the Export  
button in the MS Method editor. For more  
information on exporting a .BIC file from .MSM,  
see the Mariner Workstation Users Guide.  
3. In the Save As dialog box, type a name for the exported  
file, then click Save.  
Saving .SET files To save processing and graphic settings (.SET) from a .DAT or  
results file:  
1. Open or activate the .DAT, .RSD or .RCD file.  
2. From the File menu, select Settings, then select one of  
the following:  
Save Processing Settings As  
Save Graphic Settings As  
Save Graphic/Processing Settings As  
3. In the Save As dialog box, type a name for the .SET file,  
then click OK.  
For information, see Section 1.4.2, Customizing Processing  
and Graphic Settings (.SET).  
Data ExplorerSoftware Users Guide  
1-37  
       
Chapter 1 Data ExplorerBasics  
Saving .LBS To save spectrum (.LBS) or chromatogram (.LBC) peak label  
files from a .DAT, .RSD, or .RSC file, see Section 3.5.3,  
Setting Custom Peak Labels.  
and .LBC files  
1
1.6.6 Copying from Data Files  
Overview You can copy the following types of data from data and result  
files to the Windows clipboard:  
Trace ImageCopies the graphic image of the trace in  
Trace DataCopies the raw data for the trace  
displayed in the active window. You can also use this  
command by right-clicking, then selecting Copy Trace  
Data. For more information, see Section 2.5.2, Copying  
Traces from Multiple Data Files to a Window.  
Displayed PeaksCopies peak list entries for the  
peaks displayed in the active window.  
All PeaksCopies peak list entries for all peaks in the  
active view.  
Mass ListCopies all centroid or apex masses from  
the peak list for the active Spectrum window.  
Copy trace image To copy the trace as it is displayed in the active window to the  
Windows clipboard:  
1. Select the trace window to copy. Zoom and adjust the  
trace as needed.  
2. From the Edit menu, select Copy, then select Trace  
Image.  
3. Paste the image into an application that handles Windows  
Metafile format images, for example Microsoft  
PowerPoint®.  
1-38  
Applied Biosystems  
       
Managing Files  
NOTE: If you paste the image into an application that  
does not handle Windows Metafile format images, for  
example Microsoft Paint, images are distorted.  
1
Copy trace data To copy raw data (x,y pairs) for the peaks displayed in the  
active trace to the Windows clipboard:  
1. Select the trace window to copy. Zoom and adjust the  
trace as needed.  
NOTE: Only raw data for the set of peaks displayed in  
the trace window is copied.  
2. From the Edit menu, select Copy, then select Trace Data.  
3. Paste the data into an appropriate application, for  
example, another Data Explorer trace window or Microsoft  
Excel.  
When pasted into a Data Explorer trace window, the  
trace appears with the filename in parentheses and the  
trace label from the copied trace.  
CAUTION  
A trace pasted into an active Data Explorer window  
contains only the data points for the trace. The Sample  
Info and Instrument settings tabs display data from the  
original data file. These tabs do not include information  
Copy displayed Use this method to copy the section of the peak list pertaining  
to the peaks displayed in the active trace window. To copy the  
peak list for all peaks in the active trace, see Copy all peaks”  
peaks  
on page 1-40.  
Data ExplorerSoftware Users Guide  
1-39  
           
Chapter 1 Data ExplorerBasics  
NOTE: Copy Displayed Peaks copies all fields and  
1
headings. However, some data applications may not work  
correctly if headings are present because the first row  
contains text and not data. For information on copying the  
peaks list without headings, see Section 3.3.3, Saving the  
Peak List.  
To copy the peak list for the displayed peaks to the Windows  
clipboard:  
1. Set peak detection as needed to create a peak list. See  
Section 3.2, Peak Detection.  
2. Select the trace window to copy. Zoom and adjust the  
trace as needed.  
NOTE: Only peak list information for the set of peaks  
displayed in the trace window is copied.  
3. From the Edit menu, select Copy, then select Displayed  
Peaks.  
example Microsoft Excel.  
Copy all peaks Use this method to copy the peak list for all peaks in the active  
trace. To copy only the section of the peak list pertaining to the  
peaks displayed in the active trace window, see Copy  
displayed peakson page 1-39.  
NOTE: Copy All Peaks copies all fields and headings.  
However, some data applications may not work correctly if  
headings are present because the first row contains text  
and not data. For information on copying the peaks list  
To copy the peak list for all peaks in the active trace to the  
Windows clipboard:  
1. Set peak detection as needed to create a peak list. See  
Section 3.2, Peak Detection.  
1-40  
Applied Biosystems  
     
Managing Files  
3. Display the peak list by selecting Output Window from  
the Display menu, then clicking the Chro Peak List or  
Spec Peak List tab.  
1
4. Sort the peak list as needed. See Sorting the peak list”  
on page 3-42.  
5. From the Edit menu, select Copy, then select All Peaks.  
6. Paste the data into an appropriate application, for  
example, Microsoft Excel.  
Copy mass list This function copies only the masses for the spectrum and is  
useful for pasting data directly into spreadsheets or other  
applications without having to delete header information.  
To copy all centroid or apex mass data from the spectrum  
peak list to the clipboard:  
1. Select the Spectrum window to copy from.  
2. Select Peak Label from the Peaks menu.  
3. Specify the Mass Type by selecting Apex or Centroid.  
For more information, see Section 3.5.2, Setting  
Chromatogram and Spectrum Peak Labels.  
4. Click OK.  
5. From the Edit menu, select Copy, then select Mass List.  
Hint: If you activate the Spec Peak List tab, you can  
copy the mass list by right-clicking, then selecting Copy  
Mass List from the popup menu.  
The mass list is copied to the clipboard with a header  
indicating Apex Mass or Centroid Mass.  
6. Paste the data into an appropriate application, for  
example, Microsoft Excel.  
Data ExplorerSoftware Users Guide  
1-41  
     
Chapter 1 Data ExplorerBasics  
1
1-42  
Applied Biosystems  
Chapter  
2 Using Chromatogram  
and Spectrum  
2
Windows  
This chapter contains the following sections:  
2.1  
2.2  
2.3  
2.4  
2.5  
2.6  
2.7  
Opening and Closing Data Files......................................... 2-2  
Adjusting the Display Range ............................................ 2-11  
Organizing Windows ........................................................ 2-13  
Manipulating Traces......................................................... 2-14  
Working with Multiple Data Files ...................................... 2-36  
Saving, Opening, and Deleting .DAT Results ................... 2-38  
Exporting, Opening, and Deleting .RCD  
and .RSD Results Files (Mariner Data Only) .................... 2-39  
2.8  
Saving, Opening, and Deleting .SPC Results Files  
(Mariner Data Only) ......................................................... 2-40  
Data ExplorerSoftware Users Guide  
2-1  
   
Chapter 2 Using Chromatogram and Spectrum Windows  
2.1 Opening and Closing Data Files  
This section includes:  
Opening data files  
Displaying Mariner UV traces  
Displaying Voyager chromatograms  
Viewing read-only files  
Moving between open files  
Closing data files  
2
2.1.1 Opening Data Files  
This section includes:  
Opening a recently opened file  
Opening data files with File Open  
Automatically running macros  
Opening a The File menu lists the last several opened files (up to a  
maximum of nine). To quickly open one of these files, select it  
from the list.  
recently opened  
file  
NOTE: If you access Data Explorer software from the  
Voyager Instrument Control Panel, the most recently saved  
Voyager data file is displayed in the Data Explorer window.  
This data file and other recently acquired data files are  
listed in the Data Explorer File menu the next time you open  
Data Explorer.  
Opening data files You can open up to eight data files using the File Open dialog  
box:  
with File Open  
1. Select Open from the File menu.  
The Select Files dialog box (Figure 2-1) appears.  
2-2  
Applied Biosystems  
           
Opening and Closing Data Files  
.
2
Figure 2-1 Select Files Dialog Box  
2. Click the down arrow to display the Files of type list,  
then select the file extension to display.  
3. Select up to eight data files to open, then click Add or  
Add All. Add All is not active if the number of selected  
files exceeds eight.  
NOTE: You can also select files by double-clicking the  
file name in the file list.  
The files are listed in the Files Selected box.  
Displaying 4. To display acquisition comments before opening the data  
file, right-click a file name in the top pane of the dialog  
box, select Properties, then click the Summary tab.  
acquisition  
comment  
Data ExplorerSoftware Users Guide  
2-3  
   
Chapter 2 Using Chromatogram and Spectrum Windows  
Specifying settings 5. Select a Restoring Graphics and Processing Settings  
option to apply to new files you are opening (settings are  
not applied to files in the list that are already open):  
settings used on the data.  
Use Default SettingsApplies settings from  
the appropriate default .SET file for your  
system. See Default processing and graphic  
settingson page 1-18.  
2
Graphics and Processing Settings dialog box,  
where you select the .SET file to open.  
For information on customizing a .SET file, see  
and Graphic Settings (.SET).  
6. Click Finish to open the selected files.  
The Data Explorer window displays the selected data  
files with the processing and graphic settings you  
selected (Figure 2-2 and Figure 2-3).  
2-4  
Applied Biosystems  
     
Opening and Closing Data Files  
2
Figure 2-2 Data Explorer Window with Four Mariner Data Files Open (Each .DAT  
File Displays a Chromatogram and a Spectrum Trace)  
Figure 2-3 Data Explorer Window with Four Voyager Data Files Open (Spectrum  
Traces Only Displayed by Default)  
Data ExplorerSoftware Users Guide  
2-5  
   
Chapter 2 Using Chromatogram and Spectrum Windows  
The data is labeled accordingly:  
Type of Data  
Spectrum Trace Labels  
Mariner data  
If you open a data file you previously calibrated in Data Explorer,  
all spectra in the data file are calibrated and displayed with an  
MC or AC trace label.  
If AutoSaturation Correction is turned on, all spectra in the data  
file are corrected and displayed with an ASC trace label. For  
more information, see Section 5.11, AutoSaturation Correction  
(Mariner Data Only).  
2
Voyager data  
If you open a data file you previously calibrated in Data Explorer,  
the spectrum is calibrated and displayed with an MC trace label.  
For multispectrum data files, you can display chromatograms.  
For more information, see Section 2.1.3, Displaying Voyager  
Chromatograms.  
Automatically You can set macros to automatically run when you open or  
close files. For information, see Running Macros  
Automatically When Opening and Closing Fileson page 6-45.  
running macros  
2.1.2 Displaying Mariner DAD Traces  
When you use a diode array detector (DAD) to acquire  
Mariner data, you can display the following types of  
DAD traces:  
then select DAD TAC or DAD Channel.  
DAD spectrumDouble click a DAD chromatogram.  
For more information, see Types of Mariner dataon page 4-2  
and Creating an Extracted Absorbance Chromatogram (XAC)  
(Mariner Data Only)on page 4-13.  
2-6  
Applied Biosystems  
       
Opening and Closing Data Files  
2.1.3 Displaying Voyager Chromatograms  
To display chromatograms for multispectrum  
Voyager .DAT files:  
1. Open the .DAT files as described in Opening Data  
Fileson page 2-2.  
Spectrum traces for the .DAT files are displayed.  
2. To display a Chromatogram window, click a Spectrum  
window to activate it, then select Restore  
Chromatogram from the View menu.  
2
A Chromatogram window is displayed for the data file.  
NOTE: Diode array support and UV display are disabled  
for Voyager chromatogram traces.  
2.1.4 Viewing Read-Only Files  
For quick scanning of archived data files, you can view  
read-only files using the File Open dialog box. When you open  
a read-only file, a message indicates that the file is  
read-only and prompts you to open the file. All functions that  
write information to disk are disabled when you view a  
read-only file.  
CAUTION  
Although you can process read-only files, you cannot save  
any changes you make. When you close a read-only file  
any changes you made are lost.  
Data ExplorerSoftware Users Guide  
2-7  
           
Chapter 2 Using Chromatogram and Spectrum Windows  
2.1.5 Moving Between Open Files  
You can have more than one file open at a time. You can move  
between open files in three ways:  
Select the open file to view from the Window menu  
Use tabs  
Use the Activate File dialog box  
Using tabs To move between open files using tabs, click the tab at the  
bottom of the Data Explorer window (Figure 2-4) to activate  
the file.  
2
Double-click  
an active title bar  
to maximize a window  
Click a tab to  
select an open data file  
Figure 2-4 Tabs for Open Data Files  
If the window for the tab you click is already open, clicking the  
tab activates the window.  
If the window for the tab you click is minimized, clicking the tab  
activates the title bar for the window. Double-click the active  
title bar to display the window.  
Using Activate To move between Spectrum windows of open data files using  
Activate File:  
File  
1. Select Activate File from the Windows menu.  
2. Select a file from the Current Data File list.  
2-8  
Applied Biosystems  
               
Opening and Closing Data Files  
2
Figure 2-5 Select File to Activate Dialog Box  
3. Select:  
MaximizeTo maximize the Spectrum window  
of the selected file  
ActivateTo activate the Spectrum window of  
the selected file  
Data ExplorerSoftware Users Guide  
2-9  
 
Chapter 2 Using Chromatogram and Spectrum Windows  
2.1.6 Closing Data Files  
You can close files in the following ways:  
Select Close from the File menu to close the active file.  
Select Close All Files from the File menu to close all  
files.  
In the Open File dialog box, select any open files from  
the Files Selected list, click Remove, then click Finish  
to close the selected files.  
2
Select Exit from the File menu to close all files and exit  
the software.  
If you enabled Processing History and selected the Show  
Save History option, a dialog box is displayed. Click Save or  
Purge. For more information, see Setting Processing History  
optionson page 2-23.  
2-10  
Applied Biosystems  
   
Adjusting the Display Range  
2.2 Adjusting the Display Range  
To set the display range:  
1. Click the Chromatogram or Spectrum window to  
activate it.  
2. From the Display menu, select Range.  
3. Select X Range to set the x-axis range. The scaling units  
depend on the window you are scaling:  
2
ChromatogramScales in the same units  
currently displayed in the Chromatogram  
window (Spectrum Number or Time [Mariner  
data only])  
SpectrumScales in m/z units  
4. Select Y Range to scale the y-axis range.  
The Y Axis Setup dialog box (Figure 2-6) is displayed.  
Figure 2-6 Y Axis Setup Dialog Box  
Data ExplorerSoftware Users Guide  
2-11  
                   
Chapter 2 Using Chromatogram and Spectrum Windows  
5. Set the parameters described below as needed:  
Parameter  
Scaling Mode  
Display Relative  
Description  
Autoscales the trace to the largest peak in the selected range.  
Base Peak  
Relative  
Autoscales the trace to the base peak in the entire range, not the  
selected range. Displays a right-axis label with the base peak  
intensity.  
2
NOTE: To turn off the right axis, select Graphic Options from  
the Display menu, click the Graph #1 Setup tab, then deselect  
the Show Right Y-Axis check box.  
Absolute Value  
Sets the trace to the Y Display Range you enter in the Y Display  
Range From/To boxes.  
Display Min/Max Sets the trace display to the minimum and maximum Y values.  
Minimum Absolute Max Y  
Use Limit  
Sets the minimum value for Y-axis scaling. Useful to maintain  
relative scaling of peaks and to prevent autoscaling noise to full  
scale.  
Y Display Range  
From/To  
Sets the range for scaling. If Display Relative or Base Peak  
Relative is selected, range is in percent.  
If Absolute Value is selected, range is in counts.  
6. Click OK.  
2-12  
Applied Biosystems  
   
Organizing Windows  
2.3 Organizing Windows  
Linking views Linking Chromatogram or Spectrum windows in different data  
files allows you to zoom on multiple data files.  
NOTE: When different data files are linked, zooming  
functions performed on one data file are applied to all linked  
files. Processing and peak centering functions are applied  
to the active file only.  
2
To link data files:  
1. Open the data files you want to link.  
2. In the first data file, click the window (Chromatogram or  
Spectrum) that you want to link to another data file, then  
select Link View from the View menu.  
NOTE: Clicking  
in the toolbar links traces, not  
views.  
3. Repeat step 2 if you want to link both windows.  
4. Repeat step 2 and step 3 for the remaining data files.  
NOTE: You must select Link View for each window and  
each data file you want to link.  
Organizing  
windows  
You can organize all open windows by clicking buttons in the  
toolbar:  
Tile Horizontal  
Tile Vertical  
Cascade Windows  
You can also move and resize windows by click-dragging.  
Data ExplorerSoftware Users Guide  
2-13  
               
Chapter 2 Using Chromatogram and Spectrum Windows  
2.4 Manipulating Traces  
This section includes:  
Zooming centering, and customizing a trace  
Duplicating a trace  
Dividing the active trace  
Adding traces from the same data file to a window  
Removing traces  
Expanding and linking traces  
Recalling and rearranging traces (Processing History)  
Overlaying traces  
Annotating traces  
Viewing trace labels  
Printing traces  
2
2.4.1 Zooming, Centering,  
and Customizing a Trace  
Zooming and  
unzooming  
You can expand (zoom) an area of a trace by click-dragging a  
box around the area of interest.  
Zoom in  
NOTE: Display data cursors, then click the point you  
want to zoom on before clicking this button. See  
Setting data cursorson page 1-27.  
Zoom out to the previous zoom  
Full Unzoom  
2-14  
Applied Biosystems  
       
Manipulating Traces  
Centering a peak To center a peak in the trace window:  
in the trace  
1. Display the trace containing the peak of interest.  
2. Click the Spec Peak List or Chro Peak List tab in the  
Output window.  
3. Double-click the peak to center.  
The peak is centered in the active trace.  
Customizing a See Section 1.5, Setting Graphic Options, for information on  
customizing trace parameters such as line widths and data  
compression.  
trace  
2
2.4.2 Duplicating a Trace  
To make a duplicate of a trace:  
1. Click the trace.  
2. From the Display menu, select Duplicate Active Trace.  
The active trace is duplicated and displayed in another  
trace position.  
2.4.3 Dividing the Active Trace  
You can divide the active trace into equal segments with one  
command. This is useful to expand the x-axis range to better  
see spectral and chromatographic features.  
To divide a trace:  
1. Click the trace.  
2. From the Display menu, select Divide Active Trace, then  
select the number of segments to divide the trace into.  
The active trace is automatically split into the selected  
number of traces, with the range evenly divided among  
the traces.  
Data ExplorerSoftware Users Guide  
2-15  
         
Chapter 2 Using Chromatogram and Spectrum Windows  
For example, if you select Divide Active Trace to Four  
when the active trace has a range of 0.0 to 20.0, the  
active trace is divided into four traces:  
First trace represents the range from 0 to 5  
Second trace represents the range from 4 to 10  
Third trace represents the range from 9 to 16  
Fourth trace represents the range from 15 to 20  
NOTE: To restore the display to a single trace, select  
Remove Inactive Traces from the Display menu.  
2
2.4.4 Adding Traces from  
the Same Data File to a Window  
This section describes:  
Overview  
Setting the Replace mode  
Adding a trace  
Overview  
By default, the Data Explorer software displays:  
Data Type  
Mariner  
Window  
Chromatogram and Spectrum.  
Voyager single Spectrum only.  
spectrum  
Voyager  
multispectrum  
Spectrum only. Can also optionally  
display chromatogram.  
Chromatogram and Spectrum windows each contain  
one trace (see Figure 1-3 on page 1-11).  
2-16  
Applied Biosystems  
       
Manipulating Traces  
When you perform certain functions (for example,  
smoothing), a new trace is created. You can set the Replace  
mode to add to or replace the active trace.  
You can add up to seven new traces to a window to allow you  
to keep original data displayed when you generate new  
traces.  
Setting the To set the Replace mode:  
Replace mode  
1. From the Display menu, select Add/Remove Traces.  
The Display Trace dialog box is displayed (Figure 2-7).  
2
Figure 2-7 Display Trace Dialog Box  
Data ExplorerSoftware Users Guide  
2-17  
       
Chapter 2 Using Chromatogram and Spectrum Windows  
2. Select the Replace Mode:  
Replace the Active Trace (default)Replaces  
the active trace with the newly created trace.  
Add a New TraceAdds the newly created  
trace to the window. The original trace remains  
displayed and accessible, allowing visual  
comparison of the traces.  
2
NOTE: If Replace Mode is set to Add a New Trace and  
eight traces are present when you perform a function,  
the active trace is replaced when a new trace is  
Hint: A toolbar button is available for switching  
between Replace and Add mode. See Customizing  
toolbarson page 1-21 for information. The  
button  
is located in the Graph category.  
3. If desired, you can also add traces by selecting the trace  
to add from the Select Traces to Display section.  
4. Click OK.  
Adding a trace To add a new trace to a window:  
1. Click the Chromatogram or Spectrum window.  
2. Click in the toolbar.  
NOTE: You can also add traces by selecting Add/  
Remove Traces from the Display menu.  
The software adds a trace to the window. You can add  
up to seven traces (Figure 2-8).  
Hint: Resize the window to view all added traces.  
2-18  
Applied Biosystems  
             
Manipulating Traces  
Original  
trace  
Added  
traces  
2
Figure 2-8 Adding Traces (Four Traces Shown, up to Four More Can Be Added)  
When you perform a function that adds a new trace, the label  
of the trace changes from Not Used to the label for the type of  
trace created (Figure 2-9).  
Data ExplorerSoftware Users Guide  
2-19  
 
Chapter 2 Using Chromatogram and Spectrum Windows  
Figure 2-9 shows the original trace and three added traces  
that now contain a smoothed spectrum (SM), a centroided  
spectrum (CT), and a baseline offset spectrum (BO).  
Original  
trace  
2
Added  
traces  
Figure 2-9 Added Traces Containing Data  
For a description of trace labels, see Viewing Trace Labels”  
on page 2-30.  
2-20  
Applied Biosystems  
 
Manipulating Traces  
2.4.5 Removing Traces  
Removing the To remove the active trace from a window:  
active trace  
1. Click the trace to remove.  
2. Click in the toolbar, or right-click the trace, then  
select Remove Trace from the menu.  
The trace is removed.  
Removing To quickly remove all inactive traces from the window:  
2
inactive traces  
1. Click the trace to keep displayed to make it the active  
trace.  
2. Select Remove Inactive Traces from the Display menu.  
2.4.6 Expanding and Linking Traces  
When you have more than one trace displayed in a window,  
you can:  
Expand tracesClick a trace, then click  
toolbar to expand the selected trace for closer  
examination. To display all traces, click  
in the  
again.  
Link tracesClick  
in the toolbar to link all traces.  
Any zooming actions you perform on one trace affect all  
traces in that window. To unlink traces, click  
again.  
Data ExplorerSoftware Users Guide  
2-21  
         
Chapter 2 Using Chromatogram and Spectrum Windows  
2.4.7 Recalling and Rearranging Traces  
(Processing History)  
Overview The Chromatogram and Spectrum windows can display up to  
8 traces at a time for a data file. However, up to 16 traces are  
held in memory.  
You can recall traces previously displayed for an open data  
file, or rearrange the order of traces, using the Processing  
History command. You can also set Processing History  
Options to automatically delete history or to disable the  
function completely.  
2
Recalling or To recall or rearrange previously displayed traces:  
rearranging  
1. Click the trace position in which you want to recall or  
rearrange a trace.  
2. From the Display menu, select Processing History.  
A submenu is displayed, listing the last 16 processed  
traces viewed in the window.  
NOTE: Unprocessed traces and theoretical traces  
generated using commands on the Applications menu  
are not listed.  
3. Select the trace to display.  
The selected trace is displayed in the active trace  
position.  
Removing traces To remove traces from the Processing History list:  
from the history  
list  
1. From the Display menu, select Processing History.  
2. Select Remove History.  
3. Select the traces to remove from the list, then click OK.  
2-22  
Applied Biosystems  
       
Manipulating Traces  
Setting To set Processing History options:  
Processing  
History options  
1. From the Tools menu, select Processing History  
Options.  
The Processing History Options dialog box  
(Figure 2-10) is displayed.  
2
Figure 2-10 Processing History Options Dialog Box  
2. Turn Processing History on or off. If you turn on, specify:  
Purge processing historyRecords all  
processing functions performed. Stores the  
history log in a .CTS file and purges the history  
log when you close the data file.  
Purging a history file does not affect the data  
contained in the data file. It clears the contents  
of the .CTS file that contains the processing  
history.  
Save processing historyRecords all  
processing functions performed. Stores the  
history log in a .CTS file and maintains the  
history log when you close the data file.  
If you save processing history, .CTS files can  
become very large. You can periodically delete  
older .CTS files to clear disk space.  
3. If you want a reminder dialog box to appear when you  
close a data file that prompts you to selectively save or  
purge the history file, select Show Save History dialog.  
4. Click OK.  
Data ExplorerSoftware Users Guide  
2-23  
       
Chapter 2 Using Chromatogram and Spectrum Windows  
2.4.8 Overlaying Traces  
This section includes:  
Overlaying traces from different data files  
Overlaying traces in a single data file  
Changing the active trace  
Overlaying traces To overlay traces from different data files:  
2
from different  
data files  
1. Copy chromatogram or spectrum traces you want to  
overlay into a trace window. For more information, see  
Section 2.5.2, Copying Traces from Multiple Data Files  
to a Window.  
Hint: The copied traces display the original trace label  
and filename.  
2. To use settings other than defaults, set attributes for  
the overlay if needed. See Setting overlay attributes”  
on page 2-26.  
3. Click the trace of interest to activate it.  
NOTE: Only the active trace in an overlay is affected by  
processing tools. However, all traces are affected by  
zooming functions.  
4. From the Display menu, select Overlay Traces.  
Hint: A toolbar button is available to toggle between  
Overlay and Undo Overlay mode. See Customizing  
toolbarson page 1-21 for information. The  
button  
is located in the Graph category.  
The traces are overlaid. The trace names are listed in  
the trace label.  
2-24  
Applied Biosystems  
       
Manipulating Traces  
NOTE: When saving results, only the results for the  
Overlaying traces To overlay traces in a single data file:  
in a single data  
file  
1. Display the chromatogram or spectrum traces you want  
to overlay. For more information, see Section 2.4.4,  
Adding Traces from the Same Data File to a Window  
2. To use settings other than defaults, set attributes for the  
overlay if needed. See Setting overlay attributeson  
page 2-26.  
2
3. Click the trace of interest to activate it.  
NOTE: Only the active trace in an overlay is affected by  
processing tools. However, all traces are affected by  
zooming functions.  
4. From the Display menu, select Overlay Traces.  
Hint: A toolbar button is available to toggle between  
Overlay and Undo Overlay mode. See Customizing  
toolbarson page 1-21 for information. The  
button  
is located in the Graph category.  
The traces are overlaid. The trace names are listed in  
the trace label.  
NOTE: When saving results, only the results for the  
active trace are saved.  
Data ExplorerSoftware Users Guide  
2-25  
 
Chapter 2 Using Chromatogram and Spectrum Windows  
Changing the To change the active trace in an overlay:  
active trace  
1. From the Display menu, deselect Overlay.  
2. Click the trace to activate.  
3. From the Display menu, select Overlay.  
Sequentially When you display overlaid traces, you can sequentially  
activate each trace by clicking  
and  
.
activating  
overlaid spectra  
NOTE: If you overlay processed spectra, and then display a  
different spectrum number, the processed spectrum is lost.  
For example, assume you smooth Spectrum #3 in a trace,  
then overlay it with another trace. If you select Spectrum  
2
#3, then click  
to display Spectrum #2, then click  
to return to Spectrum #3, Spectrum #3 is no longer  
smoothed.  
Setting overlay To set overlay attributes:  
attributes  
1. Display the individual traces to overlay.  
2. Select Graphic Options from the Display menu.  
The Graph and Plot Options dialog box is displayed.  
NOTE: If traces are overlaid when you select Graphic  
Options, you can set attributes for the active trace only.  
3. Click View Setup.  
4. To apply the graphic options to all traces, select the Use  
same settings for all traces check box.  
NOTE: If the Use same settings for all traces check box  
is selected when traces are overlaid, the attributes are  
applied to the active trace only. When you display  
individual traces again, settings are applied to all  
traces.  
2-26  
Applied Biosystems  
     
Manipulating Traces  
NOTE: You must select the Use same settings for all  
traces check box before selecting options for traces. If  
you do not, settings are applied only to the active trace.  
5. In View Setup, select Overlay Trace scaling:  
Display RelativeAutoscales each trace to the  
base peak in the display range  
BP RelativeAutoscales each trace to the base  
2
peak in the trace  
Absolute ValueMaintains scaling of individual  
traces  
6. Set trace colors as needed:  
If you selected Use same settings for all traces,  
click Autocolor to allow the software to  
automatically assign trace colors.  
NOTE: The active trace color stays at its original  
setting. Other trace colors are set based on the  
active trace color. For example, if the active trace  
is yellow, other traces are assigned the colors pale  
blue, pale green and medium gray, which are the  
colors listed after yellow in the Trace color list,  
because white may not print on certain printers.  
If you did not select Use same settings for all  
traces, click Graph Setup, then select trace  
colors. See Setting colorson page 1-25 for  
information.  
7. Click OK.  
Data ExplorerSoftware Users Guide  
2-27  
       
Chapter 2 Using Chromatogram and Spectrum Windows  
2.4.9 Annotating Traces  
You can add text annotation to traces by:  
Copying a line of results from the Output window (or  
copying any ASCII text from any source), then pasting  
the copied information on the trace  
Typing text on the trace  
Copying results To copy results:  
2
information, see:  
Section 5.3, Manual Calibration, or Section 5.4,  
Automatic Calibration  
Section 5.6, Mass Deconvolution (Mariner Data  
Only)  
Section 6.2, Using the Isotope Calculator  
Section 6.3, Using the  
Mass Resolution Calculator  
Section 6.4, Using the Signal-to-Noise Ratio  
Calculator  
2. Click the Results tab in the Output window.  
3. Select the line of text to copy, then right-click and select  
Copy from the menu displayed.  
NOTE: If you select more than one line of text, only the  
first line is pasted when you annotate the trace.  
2-28  
Applied Biosystems  
         
Manipulating Traces  
Annotating the To annotate the trace:  
trace  
1. Click the trace at the location where you want to insert  
text.  
2. Right-click, then select:  
Paste textIf you copied results.  
Add text annotationIf you want to type in  
text. Type in text as needed.  
The text is added to the trace, and remains in the  
Spectrum window until you delete it.  
2
NOTE: The text is associated with the x-coordinate. If  
you display another spectrum, the text remains in view.  
If you zoom on a different region of the trace and the x-  
coordinate moves out of view, the annotated text also  
moves out of view. If you annotate overlaid traces, only  
the text associated with the active trace is displayed.  
NOTE: To move the text, click-drag the text to the  
desired position.  
3. To customize the appearance of the annotated text, see  
Section 1.5, Setting Graphic Options.  
4. To delete an annotation, do either of the following:  
Select the text, right-click, then select the  
appropriate delete or cut option.  
Right-click the trace, then select Delete All  
Text.  
NOTE: Text annotations are included when you print  
the spectrum. They are not saved with the spectrum  
when you close the data file.  
Data ExplorerSoftware Users Guide  
2-29  
       
Chapter 2 Using Chromatogram and Spectrum Windows  
2.4.10 Viewing Trace Labels  
The Data Explorer software includes a label in the trace  
header to identify the type of data displayed.  
NOTE: Trace labels are applied by the software and cannot  
be removed.  
Chromatogram Labels in the chromatogram title identify the following types of  
displayed data:  
2
trace labels  
Chromatogram Trace  
Description  
Label  
BC  
Baseline corrected  
BO  
Baseline offset  
BP  
Base Peak mass and intensity  
Constant neutral loss chromatogram  
Event filtered chromatogram (MS Method data only)  
CNL  
EF  
Mass xxx:yyy  
Extracted ion chromatogram for a mass range where xxx  
is the starting mass and yyy is the ending mass  
NFX  
NRX  
SMX  
Noise-filtered trace, where X is the applied Correlation  
Factor  
Noise-removed trace, where X is the number of standard  
deviations of noise removed  
Smoothed trace, where X is the number of smoothing  
points applied  
TAC  
Total absorbance chromatogram  
+ TAC  
XAC  
Realigned total absorbance chromatogram  
Extracted absorbance chromatogram  
2-30  
Applied Biosystems  
                               
Manipulating Traces  
Chromatogram Trace  
Label  
Description  
+ XAC  
Realigned extracted absorbance chromatogram  
XIC xxx ± yyy  
Extracted ion chromatogram for a selected mass where  
xxx is the center mass and yyy is the specified window  
NOTE: Extracted ion chromatograms were previously  
labeled with Massinstead of XIC.  
2
Figure 2-11 illustrates an extracted ion chromatogram with a  
XICchromatogram trace label.  
Trace label  
Figure 2-11 Chromatogram Trace Label  
Spectrum trace Labels in the spectrum title identify the following types of data  
displayed:  
labels  
Spectrum Trace Label  
Type of Processing  
+, /, or mmm:nnn  
Added, accumulated, or subtracted spectrum from  
(Mariner data only)  
spectrum #mmm to #nnn  
AC  
Automatically mass calibrated  
Advanced baseline correction  
AdvBC  
ASC (Mariner data only) AutoSaturation corrected trace  
BC  
BO  
Baseline corrected  
Baseline offset  
Data ExplorerSoftware Users Guide  
2-31  
                           
Chapter 2 Using Chromatogram and Spectrum Windows  
Spectrum Trace Label  
Type of Processing  
BPI  
CT  
Base peak intensity  
Centroid  
DAD  
Diode array data  
DECONV  
Zero charge deconvoluted trace  
(Mariner data only)  
2
DI  
Deisotoped trace  
Isotope  
ISO  
MC  
NF  
Manually mass calibrated  
Noise-filtered trace  
NRX  
Noise-removed trace, where X is the number of standard  
deviations of noise removed  
RSMX (does not apply  
to Voyager PSD data)  
Default smoothing applied, where X is the resolution  
value (from peak detection) used to calculate the  
optimum number of smoothing points to apply at every  
mass point  
SC (Mariner data only)  
SMX  
Converted to single-charge  
Smoothed, where X is the number of smoothing points  
applied  
Stitched PSD  
TR  
Composite PSD spectrum  
Truncated spectrum  
2-32  
Applied Biosystems  
                           
Manipulating Traces  
Figure 2-12 illustrates a smoothed spectrum with an SM5”  
spectrum trace label.  
Trace label  
Figure 2-12 Spectrum Trace Label  
2
2.4.11 Printing Traces  
Printing traces To print traces:  
1. Display the traces to print.  
To obtain a clear printout, you can set the Trace Color  
and other attributes to dark colors before printing traces  
selecting White Background.  
NOTE: If you previously modified the colors associated  
with this command as described in Section 1.5.1,  
Changing Background Color, selecting this command  
may not set a white background and black traces.  
When you manually set colors, note the following:  
Selections set to white (or line widths set to 0)  
may not print on certain printers.  
If you select different trace colors for multiple  
traces, only the color for the active trace is saved.  
2. Click the window (Chromatogram or Spectrum) to print.  
Data ExplorerSoftware Users Guide  
2-33  
               
Chapter 2 Using Chromatogram and Spectrum Windows  
3. To print with the x-axis along the longest length of the  
paper, select Print Setup from the File menu, then select  
NOTE: The Landscape printing orientation you set in  
Data Explorer is lost when you close Data Explorer. To  
permanently set the printer, see Dedicating a printer to  
landscape orientationon page 2-35.  
4. From the File menu, select a print option:  
2
Print File with Instrument SettingsPrints  
displayed trace windows from the active file  
along with the settings used to obtain the data.  
Print Spectrum or Chromatogram Trace—  
Prints the active trace.  
Print Spectrum or Chromatogram View—  
Prints all traces in the Spectrum or  
Chromatogram window.  
Print All ViewsPrints traces for all open data  
files.  
NOTE: If you select Print All Views when more  
than two data files are open, certain printers may  
not print the data file name. To ensure data file  
names are printed, print views individually, or only  
open two data files before you select Print All  
Views.  
NOTE: To print a trace that is displayed as Vertical  
Bars, change the Line Width to 1. If Line Width is set to  
0, Vertical Bar traces may not print on certain printers.  
See Section 1.4, Customizing the Data Explorer  
Window, for information.  
2-34  
Applied Biosystems  
           
Manipulating Traces  
NOTE: Line Widths of 0 or 1 (or lines set to the color  
white) may not print on certain printers. If traces do not  
print, change the line width (or color).  
Dedicating a To dedicate the printer to landscape orientation:  
printer to  
landscape  
orientation  
1. From the Windows desktop, click Start, then select  
Settings.  
2. Click Printers.  
2
3. Select the printer name in the displayed list.  
4. Click File, then select Document Defaults.  
5. In the Page Setup tab, select Landscape orientation.  
NOTE: If you cannot select Landscape orientation, you  
do not have access permission. See your administrator.  
Print Setup The Print Setup function allows you to select a printer and set  
printer options. For more information on Print Setup and on  
connecting printers to your computer, refer to the  
documentation provided with your computer.  
Data ExplorerSoftware Users Guide  
2-35  
       
Chapter 2 Using Chromatogram and Spectrum Windows  
2.5 Working with Multiple Data Files  
When you have multiple data files open, you can:  
Work with the data files separately to view, zoom, and  
print  
Copy traces from one data file to another to compare or  
combine data  
2
To view, zoom, and print multiple data files:  
1. Open the data files as described in Section 2.1,  
Opening and Closing Data Files.  
2. Set the Y Scaling mode to Display Relative, as described  
in Section 2.2, Adjusting the Display Range.  
3. Link the data files by selecting the Spectrum window for  
each data file, then selecting Link View from the View  
menu.  
You must select Link View for each data file.  
NOTE: Clicking  
in the toolbar links traces, not  
views.  
4. Organize the Spectrum windows for the data files by  
clicking and in the toolbar.  
5. Zoom and manipulate traces as needed.  
6. To print all data files, select Print All Views from the File  
menu.  
2-36  
Applied Biosystems  
             
Working with Multiple Data Files  
NOTE: If you select Print All Views when more  
than two data files are open, certain printers may  
not print the data file name. To ensure data file  
names are printed, print views individually, or only  
open two data files before you select Print All  
Views.  
2.5.2 Copying Traces from  
2
Multiple Data Files to a Window  
You can copy up to seven traces from open data files to a  
different trace window:  
1. Select the trace to copy.  
2. From the Edit menu, select Copy, then Trace Data, or  
right-click the trace, then select Copy Trace Data.  
3. Activate the window in which to paste the trace.  
NOTE: The Add/Replace Trace state, as set in the  
Display Trace dialog box, determines whether the  
copied trace replaces or is added below the active  
trace. See Setting the Replace modeon page 2-17.  
4. From the Edit menu, select Paste, then Trace Data, or  
right-click the window, then select Paste Trace Data.  
The copied trace is added to the active window. The  
original trace label is preceded by the name of the file  
from which you copied the trace.  
You can repeat step 1 through step 4 until a maximum of eight  
traces are displayed in a trace window. If eight traces are  
displayed and you copy a new trace to the window, the active  
trace is replaced.  
Data ExplorerSoftware Users Guide  
2-37  
           
Chapter 2 Using Chromatogram and Spectrum Windows  
Comparing After you copy a trace to another trace window, you can  
compare traces by overlaying (see Section 2.4.8, Overlaying  
copied traces  
Traces) or by using trace arithmetic (see Section 5.12, Adding  
and Subtracting Raw or Processed Spectra from the Same or  
Different Data Files (Dual Spectral Trace Arithmetic).  
2.6 Saving, Opening, and  
Deleting .DAT Results  
2
Saving results for To save results for .DAT files:  
.DAT files  
1. Process a data file to generate results as needed.  
2. From the File menu, select Result Spectrum or Result  
Chromatogram, then select Save As.  
3. Use the default title, or enter a name (up to 31 characters)  
for the results in the Title text box, then click OK.  
Only the results for the active trace are saved.  
NOTE: Results are stored within the .DAT file, not as  
separate files. Title is an identifier you can use to recall  
the results. It is not a file name.  
Opening results To open results for .DAT files:  
for .DAT files  
1. From the File menu, select Result Spectrum or Result  
Chromatogram, then select Open.  
2. Select the title of the previously saved results to open,  
then click OK.  
3. Click the Sample Info tab in the Output window to display  
information for the result.  
Deleting results To delete results for .DAT files:  
for .DAT files  
1. From the File menu, select Result Spectrum or Result  
Chromatogram, then select Delete.  
2. Select the title of the results to delete, then click OK.  
2-38  
Applied Biosystems  
                 
Exporting, Opening, and Deleting .RCD and .RSD Results Files (Mariner Data Only)  
2.7 Exporting, Opening, and  
Deleting .RCD and .RSD Results  
Files (Mariner Data Only)  
Exporting results To export results to a .RSD or .RCD file:  
for .RCD and  
.RSD files  
1. With a data file open in Data Explorer, select a  
spectrum or chromatogram.  
2
2. From the File menu, select Export, then Result  
Chromatogram or Result Spectrum.  
A Save As dialog box appears.  
3. Specify the name and destination of the exported file. By  
default, the software assigns a .RCD extension for a  
chromatogram or a .RSD extension for a spectrum.  
4. Click OK.  
Opening results To open results from .RCD and .RSD files:  
for .RCD and  
.RSD files  
1. Select Open from the File menu.  
The Select Data File to Open dialog box is displayed.  
2. From the Files of Type drop-down list, select Result  
Spectrum Files (*.RS*) or Result Chromatogram Files  
(*.RC*).  
3. Select the .RCD or .RSD file to open, then click OK.  
4. Click the Sample Info tab in the Output window to display  
the following information for the result file:  
Name of the original raw data file the result file  
was generated from  
Processing functions that were performed and  
saved in the result file  
Data ExplorerSoftware Users Guide  
2-39  
               
Chapter 2 Using Chromatogram and Spectrum Windows  
Deleting results Use Windows NT Explorer to delete .RCD and .RSD result  
files.  
for .RCD and .RSD  
files  
2.8 Saving, Opening, and Deleting  
.SPC Results Files  
(Mariner Data Only)  
2
Saving results for To save results for .SPC files (Mariner data only):  
.SPC files  
1. Process a data file to generate results as needed.  
2. From the File menu, select Result Spectrum or Result  
Chromatogram, then select Save As.  
3. Enter a file name for the result file in the File Name text  
box.  
Spectrum files are automatically named with an .RST  
extension. Chromatogram files are automatically  
named with an .RCT extension.  
4. Click OK.  
NOTE: You can also save Mariner .RCT and .RST files  
from .DAT format data files in the Instrument Control  
Panel. For more information, see the Mariner  
Workstation Users Guide.  
Opening results: To open results for .SPC files:  
for .SPC files  
1. Select Open from the File menu.  
The Select Data File to Open dialog box is displayed.  
2. From the Files of Type drop-down list at the bottom of the  
dialog box, select Result Spec Files (*.RS*) or Result  
Chro Files (*.RC*).  
2-40  
Applied Biosystems  
                     
Saving, Opening, and Deleting .SPC Results Files (Mariner Data Only)  
3. Select the .RST or .RCT file to open, then click OK.  
NOTE: Saturation Correction is not applied to Mariner  
.RST files. For more information, see Section 5.11,  
AutoSaturation Correction (Mariner Data Only).  
4. Click the Sample Info tab in the Output window to display  
the following information for the result file:  
Name of the original raw data file from which the  
2
result file was generated  
Processing functions that were performed and  
saved in the result file  
Deleting results Use Windows NT Explorer to delete .RST and .RCT result files  
generated from .SPC files.  
for .SPC files  
Data ExplorerSoftware Users Guide  
2-41  
 
Chapter 2 Using Chromatogram and Spectrum Windows  
2
2-42  
Applied Biosystems  
Chapter  
3
3 Peak Detection  
and Labeling  
This chapter contains the following sections:  
3.1  
3.2  
3.3  
3.4  
3.5  
3.6  
Overview ......................................................... 3-2  
Peak Detection ................................................ 3-6  
Peak List ....................................................... 3-37  
Deisotoping a Spectrum................................. 3-45  
Peak Labeling................................................ 3-52  
Process that Occurs During Peak Detection,  
Centroiding, and Integration .......................... 3-67  
3.7  
Default Peak Detection Settings .................... 3-71  
Data ExplorerSoftware Users Guide  
3-1  
     
Chapter 3 Peak Detection and Labeling  
3.1 Overview  
This section includes:  
Default peak detection  
The resolution-based peak detection routine  
3.1.1 Default Peak Detection  
Overview When you open a data file, it is automatically peak detected.  
For:  
Chromatographic dataThe software uses default  
settings that have been optimized to yield acceptable  
peak detection for many sample types. For more  
information, see Section 3.7, Default Peak Detection  
Settings.  
3
Spectral dataThe software uses a resolution-based  
peak detection routine to calculate peak detection values  
that provide optimum peak detection for most sample  
types. For more information, see Section 3.1.2, The  
Resolution-Based Peak Detection Routine.  
For many applications, the default peak detection settings and  
settings calculated by the resolution-based peak detection  
routine provide acceptable peak detection. If default settings  
do not provide acceptable peak detection, you can adjust the  
settings as described in:  
Section 3.2.2, Strategy for Voyager Peak Detection  
NOTE: If peak detection settings do not detect desired  
peaks, you can manually insert peaks. See Section 3.3.2,  
Inserting Peaks in the Peak List.  
3-2  
Applied Biosystems  
           
Overview  
3.1.2 The Resolution-Based  
Peak Detection Routine  
This section describes:  
Type of data affected  
Process that occurs  
PSD peak detection for Voyager data  
Type of data The resolution-based peak detection routine applies to  
spectral data only. You can enable or disable the  
resolution-based peak detection routine as described on  
page 3-14.  
affected  
Process that When enabled, the resolution-based peak detection routine  
automatically:  
occurs  
Divides a trace into detection ranges based on the  
expected number of data points across a typical mass  
spectral peak.  
3
Applies a Filter Width that is equal to the number of data  
points per peak in each detection range.  
Uses a Filter Increment of 1.  
Data ExplorerSoftware Users Guide  
3-3  
           
Chapter 3 Peak Detection and Labeling  
The software uses the following formula to calculate the  
expected number of data points in a peak:  
Flight time  
to which the  
Expected  
number of  
data points  
=
data point corresponds  
(2 × mass resolution × Bin size)  
Where:  
Mass resolution is a user-defined value.  
Data-type-dependent defaults are provided, but can be overwritten.  
Bin size is an instrumental constant:  
· For Mariner data, Bin size = 1 ns.  
· For Voyager data, Bin size is variable based on  
the digitizer used and the mass range acquired.  
3
Detection ranges Figure 3-1 is an example of the resolution-based detection  
ranges automatically calculated by the software.  
Number  
of data points  
41 47 53 59 65  
71  
77  
83  
Figure 3-1 Example Detection Ranges Calculated by Software Based on Number  
of Data Points Per Peak  
3-4  
Applied Biosystems  
       
Overview  
Overlapping peak To accommodate spectral peaks that occur on the boundary of  
two peak detection ranges, the software creates detection  
ranges that overlap (Figure 3-2).  
detection ranges  
Peak A  
Peak B  
Peak C  
Peak D  
Range 4  
Range 3  
Range 2  
Range 1  
Figure 3-2 Overlapping Detection Ranges  
3
If a peak occurs in an overlapping range (for example, Peak B  
in Figure 3-2), the software detects the peak using the settings  
from the detection range within which the entire peak occurs  
(Range 2 in Figure 3-2).  
If a peak occurs completely within two ranges (for example,  
Peak D in Figure 3-2), the software uses the settings from the  
higher m/z detection region (Range 4 in Figure 3-2).  
PSD peak For PSD data, the mass-time correlation is different from that  
of standard TOF spectra. Therefore, resolution-dependent  
settings are not supported for PSD data.  
detection for  
Voyager data  
Data ExplorerSoftware Users Guide  
3-5  
       
Chapter 3 Peak Detection and Labeling  
3.2 Peak Detection  
This section includes:  
Strategy for Mariner peak detection  
Strategy for Voyager peak detection  
Setting peak detection parameters  
Peak detection parameter descriptions  
Charge state determination and examples  
This section gives some quick suggestions on how to  
approach Mariner peak detection. For details on peak  
detection, see Section 3.2.3, Setting Peak Detection  
Parameters.  
Default peak detection values are listed in Section 3.7, Default  
Peak Detection Settings.  
3
Strategy When detecting peaks in Mariner data:  
1. Open the data file and observe the effects of the default  
peak detection settings.  
If you are analyzing resolved isotopes, default settings  
should yield acceptable peak detection. The default  
resolution setting of 5,000 is optimized for masses  
below 3,000 Da.  
2. If you are analyzing proteins, decrease the Mass  
Resolution setting.  
3-6  
Applied Biosystems  
                 
Peak Detection  
3. If peak detection is not acceptable, leave the Use  
Resolution Dependent Settings option enabled, and  
adjust the following global threshold parameters in the  
following order:  
%Base Peak IntensityUse this parameter to  
eliminate peaks with an intensity below the  
specified threshold. This threshold is represented  
as a percentage of the intensity of the base peak.  
%Max Peak AreaUse this parameter to  
fine-tune noise discrimination on the baseline or  
noise spikes on peaks. This threshold is  
represented as a percentage of the area of the  
peak with the largest area and is calculated above  
the local baseline.  
To determine an appropriate threshold, display the  
Spectrum Peak list, note the area of the peak with  
the greatest area and the areas of unwanted  
peaks, and estimate the percentage to enter as  
the threshold.  
3
Specifying a %Max Peak Area Threshold is  
particularly useful if the spectrum includes a rising  
Threshold and eliminate peaks of interest in one  
portion of the trace.  
4. If peak detection is still not acceptable, adjust the  
remaining peak detection parameters, as described in  
Section 3.2.3, Setting Peak Detection Parameters.  
Data ExplorerSoftware Users Guide  
3-7  
     
Chapter 3 Peak Detection and Labeling  
This section gives some quick suggestions on how to  
approach Voyager peak detection. For details on peak  
detection, see Section 3.2.3, Setting Peak Detection  
Parameters.  
Default peak detection values are listed in Section 3.7, Default  
Peak Detection Settings.  
Strategy When detecting peaks in Voyager data:  
1. Open the data file and observe the effects of the default  
resolution-dependent settings.  
resolution on a tall, resolved peak in the middle of the  
spectrum. If the resolution result differs by more than  
50% from the default resolution setting, fine-tune the  
Resolution setting and reapply peak detection. For  
more information, see Section 6.3, Using the  
Mass Resolution Calculator.  
3
2. To aid in peak interpretation, do all of the following:  
Baseline-correctThe Centroid% peak detection  
value is derived from a percentage of the peak  
height, which is measured from 0. For information,  
see Section 5.8.2, Using Baseline Correction.  
Noise-filter or smoothUse the method  
appropriate for your data to remove noise spikes:  
Data Type  
Linear  
Reflector  
Method  
Default or Gaussian smoothing  
Noise Reduction  
For more information, see Section 5.7, Noise  
Filtering/Smoothing.  
3-8  
Applied Biosystems  
           
Peak Detection  
Deisotope (reflector data only)Peak  
deisotoping reduces the spectrum to a  
monoisotopic centroided plot of the monoisotopic  
masses. This is useful in identifying overlapping  
isotope clusters. Make sure peak detection  
thresholds are set low enough to detect the  
monoisotopic peak before deisotoping. For more  
information, see Section 3.4, Deisotoping a  
Spectrum. Deisotoping provides no benefit on linear  
data (non-isotopically resolved) or on PSD data  
(pure isotope data).  
3. If peak detection is not acceptable, keep the Use  
Resolution Dependent Settings option enabled, and  
adjust the parameter associated with the observed  
problem:  
Problem  
Suggested Actions  
3
High-mass peaks not  
detected  
Decrease Mass Resolution setting. The default Mass  
Resolution settings are optimized for masses below  
20,000 Da.  
Noise detected as peaks Increase the %Max Peak Area.  
Decrease the Resolution.  
Peaks of interest are not Decrease the %Max Peak Area.  
detected  
NOTE: %Max Peak Area is calculated above the local  
baseline and can compensate for problems related to a  
rising global baseline.  
Low Mass Gate spike is  
identified as the Base  
Peak  
Truncate the data to eliminate the Low Mass Gate spike,  
then reapply peak detection settings. See Section 5.9,  
Truncating a Spectrum.  
Peaks detected before  
deisotoping are not  
detected after  
Set %Max Peak Area and %BP Intensity to 0 before  
deisotoping, deisotope, then reset thresholds to  
appropriate settings after deisotoping.  
deisotoping  
Data ExplorerSoftware Users Guide  
3-9  
               
Chapter 3 Peak Detection and Labeling  
Problem  
Suggested Actions  
Partially resolved peaks  
not detected  
If peaks represent two compounds, and you want both  
peaks labeled, do either of the following:  
Set %Max Peak Area to 0, then adjust the %Base  
Peak Intensity until peaks are detected.  
Click the Peak Processing tab, then change the  
default Integration Baseline Setting from  
Valley-to-Valley to Valley-to-Baseline.  
If peaks represent partially resolved isotopes, and you  
want to label and detect the average mass, decrease the  
Mass Resolution setting until the isotopic envelope is  
detected.  
4. If you see more than one of the problems listed above in a  
spectrum, you can adjust peak detection parameters for  
any or all detection ranges (referred to as setting  
parameters locally) by doing the following in the order  
listed:  
3
Disable Use Resolution Dependent Settings.  
Click the Advanced tab.  
Select a detection region that is exhibiting the  
problem.  
Click OK.  
5. If peak detection is still not acceptable, adjust the  
remaining peak detection parameters, as described in  
Section 3.2.3, Setting Peak Detection Parameters.  
3-10  
Applied Biosystems  
       
Peak Detection  
3.2.3 Setting Peak Detection Parameters  
This section includes:  
Before you begin  
Setting chromatogram parameters  
Setting Basic Settings (spectrum data)  
Setting Peak Processing parameters (spectrum data  
only)  
Resetting Basic Settings  
Before you begin Before setting peak detection, read:  
Section 3.2.1, Strategy for Mariner Peak Detection  
Section 3.2.2, Strategy for Voyager Peak Detection  
3
Setting To set chromatogram parameters for data:  
chromatogram  
parameters  
1. Click the Chromatogram window to activate it.  
2. Click the trace of interest.  
3. Click  
in the toolbar or select Peak Detection from  
the Peaks menu.  
The Chromatogram Peak Detection dialog box is  
You can click-drag the threshold cursor to adjust the  
%Base Peak Intensity.  
For a description of the parameters in the  
Chromatogram Peak Detection dialog box, see  
Chromatogram settingson page 3-19.  
Data ExplorerSoftware Users Guide  
3-11  
         
Chapter 3 Peak Detection and Labeling  
Figure 3-3 Chromatogram Peak Detection Setup  
Dialog Box  
3
4. Select a detection range and set parameters as needed.  
5. To apply settings to all traces, select Use same settings  
for all traces in view.  
To set parameters independently for all traces in a  
window, deselect Use same settings for all traces in  
view.  
6. Click Apply to accept the parameters and leave the dialog  
box open, or click OK to accept the parameters and close  
the dialog box.  
3-12  
Applied Biosystems  
   
Peak Detection  
Setting Basic Basic Settings should provide acceptable peak detection for  
most applications. After you apply these parameters, no  
further adjustment should be required. If further adjustment is  
required, select Use Advanced Settings and adjust  
parameters as needed.  
Settings  
(spectrum data)  
To set Basic Settings parameters for spectrum data:  
1. Click the Spectrum window to activate it.  
2. Click the trace of interest.  
3. Click  
in the toolbar or select Peak Detection from  
the Peaks menu.  
The Spectrum Peak Detection Setup dialog box opens,  
with the Basic Settings tab (Figure 3-4) displayed.  
NOTE: If you applied Advanced Settings to a data file,  
they override the settings on the Basic Settings tab, and  
the Advanced Settings tab is displayed when you select  
Peak Detection. See Setting Advanced Settings  
(spectrum data only)on page 3-17.  
3
4. Set Global Thresholds as needed. These thresholds are  
applied to all detection ranges, unless you override them  
for a detection range on the Advanced Settings tab.  
For a description of the parameters, see Basic Settings  
(spectrum data only)on page 3-22.  
Data ExplorerSoftware Users Guide  
3-13  
     
Chapter 3 Peak Detection and Labeling  
Figure 3-4 Spectrum Peak Detection Setup—  
Basic Settings Tab  
3
5. If you are detecting PSD data, or want to override the  
Global Thresholds, select Use Advanced Settings and  
skip to step 7.  
6. If you are detecting mass spectral data, select Use  
Resolution Dependent Settings if you want the software  
to:  
Automatically determine the number of data points  
across a peak  
Divide the trace into different detection ranges  
based on the resolution  
Apply a Filter Width and Increment appropriate for  
each detection range  
Apply a Minimum Area of 0 and a Minimum  
Intensity of 0 to all detection ranges  
3-14  
Applied Biosystems  
     
Peak Detection  
NOTE: Resolution-dependent settings do not apply to  
Mariner chromatogram data or Voyager PSD data.  
For more information, see Section 3.1.2, The  
Resolution-Based Peak Detection Routine.  
7. To apply settings to all traces, select Use same settings  
for all traces in view.  
To set parameters independently for all traces in a  
view.  
8. Type the Mass Resolution value to use for peak detection.  
The default value for the type of data displayed is  
acceptable for most applications. Defaults are listed in  
Basic Settings (spectrum data only)on page 3-22.  
3
by the Elemental Composition Calculator, the Elemental  
Targeting Application, and the Default Smoothing  
function. For more information, see Section 6.1, Using  
the Elemental Composition Calculator, Section 6.6,  
Using the Elemental Targeting Application, and  
Section 5.7, Noise Filtering/Smoothing.  
9. Click Apply to accept the parameters and leave the dialog  
box open, or click OK to accept the parameters and close  
the dialog box.  
Data ExplorerSoftware Users Guide  
3-15  
 
Chapter 3 Peak Detection and Labeling  
Setting Peak To set Peak Processing parameters:  
Processing  
parameters  
(spectrum data  
only)  
1. Click the Peak Processing tab in the Spectrum Peak  
Detection Setup dialog box.  
The Peak Processing tab is displayed (Figure 3-5).  
3
Figure 3-5 Spectrum Peak Detection Setup—  
Peak Processing Tab  
NOTE: You can enable the %BP Intensity Threshold  
Cursor and click-drag it to adjust the %Base Peak  
Intensity.  
2. Set parameters as needed, then click Apply to accept the  
parameters and leave the dialog box open, or click OK to  
accept the parameters and close the dialog box.  
For a description of the parameters, see Peak  
Processing parameters (spectrum data only)on  
page 3-26.  
3-16  
Applied Biosystems  
     
Peak Detection  
Setting Advanced To set Advanced Settings that you can apply locally to a  
selected detection range, and that override the thresholds set  
in the Basic Settings tab:  
Settings  
(spectrum data  
only)  
1. In the Basic Settings tab, select Use Advanced  
Settings.  
2. Click the Advanced Settings tab in the Spectrum Peak  
Detection Setup dialog box.  
NOTE: If you select Use Resolution Dependent  
Settings in the Basic Settings tab, Basic Settings  
override Advanced Settings. The Advanced Settings  
tab is accessible, but all parameters are dimmed. To  
make Advanced Settings available for editing, select  
Use Advanced Settings on the Basic Settings tab.  
The Advanced Settings tab is displayed (Figure 3-6).  
3
Figure 3-6 Spectrum Peak Detection Setup—  
Advanced Settings Tab  
Data ExplorerSoftware Users Guide  
3-17  
         
Chapter 3 Peak Detection and Labeling  
4. Click Apply to accept the parameters and leave the dialog  
box open, or click OK to accept the parameters and close  
the dialog box.  
For a description of the parameters and the data cursor,  
see Advanced Settings (spectrum data only)on  
page 3-28.  
Resetting Basic If you override Basic Settings by entering parameters on the  
Advanced Settings tab, you can reset Basic Settings by doing  
the following:  
Settings  
1. Click the Basic Settings tab.  
2. Select Use Resolution Dependent Settings.  
3. Enter the desired Global Thresholds.  
4. Click Apply or OK.  
The following occurs:  
3
Detection ranges, Filter Width, and Filter  
Increment previously set on the Advanced  
Settings tab are reset to defaults.  
Minimum Area and Minimum Intensity are set to 0.  
Global Threshold settings from the Basic Settings  
tab are applied to all detection ranges and  
override the thresholds set for individual detection  
ranges on the Advanced Settings tab. Detection  
ranges that do not correspond to ranges  
are deleted.  
NOTE: To reset to original peak detection settings, restore  
the default .SET file for your system. For more information,  
see Applying a .SET fileon page 1-20.  
3-18  
Applied Biosystems  
 
Peak Detection  
3.2.4 Peak Detection  
Parameter Descriptions  
This section describes:  
Chromatogram settings  
Basic Settings (spectrum data)  
Advanced Settings (spectrum data only)  
Chromatogram Table 3-1 describes the parameters in the Chromatogram  
Peak Detection Setup dialog box (see Figure 3-3 on  
page 3-12). Default peak detection values are listed in  
Section 3.7, Default Peak Detection Settings.  
settings  
Table 3-1 Chromatogram Settings  
Parameter  
Description  
3
Detection Ranges  
Specifies one or more non-contiguous m/z ranges for peak  
detection. You can set parameters for each range  
independently.  
You select a range in the Detection Ranges list box by  
single-clicking the range number.  
To add a detection range, do one of the following:  
Select an existing range, then click  
. This creates a  
new range with boundaries ranging from the end of the  
existing range to the end of the trace. If the existing  
range ends at the end of the trace, the region of the  
existing range is split in half between the existing range  
and the new range.  
Double-click an existing range to manually enter lower  
and upper boundaries.  
Select a range in the dialog box, then click-drag the  
X data cursors labels in the trace to set the lower and  
upper boundaries.  
Continued  
Data ExplorerSoftware Users Guide  
3-19  
                     
Chapter 3 Peak Detection and Labeling  
Table 3-1 Chromatogram Settings (Continued)  
Parameter  
Description  
Detection Ranges  
To delete a range, select the range, then click  
.
(continued)  
To combine all ranges in the list into one range, click  
.
The peak detection settings displayed in the dialog box  
correspond to the selected range. To view peak detection  
settings for another range, select the range of interest.  
%Base Peak  
Intensity  
Specifies a percentage of the base peak intensity as the  
threshold value.  
NOTE: This  
To be detected, peaks must be above this threshold and  
parameter was  
previously named  
Peak Threshold%  
and Base Peak  
Relative%.  
above the %Max Peak Area value.  
Hint: For most Voyager applications, leave this parameter  
set at 0 and adjust the %Max Peak Area.  
3
%Max Peak Area  
Specifies a percentage of the peak with the largest area as  
the threshold value.  
To be detected, peaks must be above this threshold and  
above the %Base Peak Intensity value.  
%Max Peak Area is calculated above the local baseline and  
baseline and signal spikes.  
Hint: If you are analyzing digest data, set %Max Peak Area  
to 0 before deisotoping to ensure that all peaks of interest  
are detected. For more information, see Section 7.2.3,  
Detecting Peaks from Complex Digests.  
Filter Width  
Specifies the number of data points used in smoothing for  
peak detection before integration. The software uses a Filter  
Width Increment of 1 when applying smoothing.  
NOTE: If you set Filter Width too high, narrow peaks may not  
be detected.  
Continued  
3-20  
Applied Biosystems  
                           
Peak Detection  
Table 3-1 Chromatogram Settings (Continued)  
Parameter  
Description  
Filter Width  
(continued)  
Hint: Set Filter Width to a number equal to the number of  
points across the peak. To determine the number of points  
across a peak, you can change the trace display from Line to  
Vertical Bars. Each vertical bar represents one data point.  
For more information, see Section 1.4, Customizing the Data  
Explorer Window.  
Integration Baseline Settings  
NOTE: You can set peak labels to display peak start, peak end, and baseline marks.  
See Section 3.5.2, Setting Chromatogram and Spectrum Peak Labels.  
Valley-to-Baseline  
Drops a vertical line from all valleys to a horizontal baseline.  
The level of the horizontal baseline is determined using the  
minimum peak valley point (left or right) for each peak. See  
Figure 3-25 on page 3-70.  
3
Valley-to-Valley  
Forces a baseline through all valley points. See Figure 3-25  
on page 3-70.  
Trace Settings  
Use same settings  
for all traces in view  
Applies settings to all traces in the active window.  
NOTE: In previous versions of Data Explorer software,  
peak detection allowed you to specify Peak Width and  
Noise Threshold for chromatogram data. The software now  
automatically:  
Uses a minimum peak width that is equal to the Filter Width,  
and a maximum peak width of 10,000 data points.  
Calculates the Noise Threshold by performing an automatic  
signal-to-noise calculation on a chromatogram and by using  
75 percent of the noise value determined as the threshold.  
Data ExplorerSoftware Users Guide  
3-21  
                     
Chapter 3 Peak Detection and Labeling  
Basic Settings Table 3-2 describes the parameters in the Basic Settings tab  
of the Spectrum Peak Detection Setup dialog box (see  
Figure 3-4 on page 3-14).  
(spectrum data  
only)  
Default peak detection values are listed in Section 3.7, Default  
Peak Detection Settings.  
Table 3-2 Basic Settings Tab Parameters (Spectrum Data Only)  
Parameter  
Description  
Global Thresholds  
%Base Peak  
Intensity  
threshold value.  
NOTE: This  
To be detected, peaks must be above this threshold and  
parameter was  
previously named  
Peak Threshold%  
and Base Peak  
Relative%.  
above the %Max Peak Area value.  
If you override Global Thresholds by selecting Use Advanced  
Settings (described on page 3-23), the software ignores  
Global Thresholds and uses the %Base Peak Intensity,  
%Max Peak Area, Minimum Area, and Minimum Intensity  
to detect peaks for the selected detection range.  
3
NOTE: If you are examining Voyager data that contains a  
Low Mass Gate spike, the software may identify the Low  
Mass Gate spike as the Base Peak. Truncate the data as  
described in Section 5.9, Truncating a Spectrum, to eliminate  
the Low Mass Gate spike and correctly identify the Base  
Peak.  
Hint: For most Voyager applications, leave this parameter  
set at 0 and adjust the %Max Peak Area.  
Continued  
3-22  
Applied Biosystems  
                       
Peak Detection  
Table 3-2 Basic Settings Tab Parameters (Spectrum Data Only) (Continued)  
Parameter  
Description  
%Max Peak Area  
Specifies a percentage of the peak with the largest area as  
the threshold value.  
To be detected, peaks must be above this threshold and  
above the %Base Peak Intensity value.  
%Max Peak Area is calculated above the local baseline and  
can compensate for problems related to a rising global  
baseline.  
If you override Global Thresholds by selecting Use Advanced  
Settings (described on page 3-23), the software ignores  
Global Thresholds and uses the %Base Peak Intensity,  
%Max Peak Area, Minimum Area, and Minimum Intensity  
thresholds set on the Advanced tab (described on page 3-28)  
to detect peaks for the selected detection range.  
3
Enable %BP  
Intensity Threshold  
Cursor  
When enabled, allows you to click-drag the cursor to set the  
% Base Peak Intensity Threshold.  
Peak Detection  
Use Resolution  
When selected, the software automatically determines  
detection ranges and uses an appropriate Filter Width and an  
Resolution-Based Peak Detection Routine.  
Dependent Settings  
(not available for  
Voyager PSD data)  
Use Advanced  
Settings  
When selected, the software ignores the Global Threshold  
values on the Basic Settings tab, and uses the threshold  
values set on the Advanced tab. See Advanced Settings  
(spectrum data only)on page 3-28.  
Continued  
Data ExplorerSoftware Users Guide  
3-23  
           
Chapter 3 Peak Detection and Labeling  
Table 3-2 Basic Settings Tab Parameters (Spectrum Data Only) (Continued)  
Parameter  
Peak Resolution  
Mass Resolution  
Value used to determine the Filter Width used for detection.  
For more information, see Section 3.1.2, The Resolution-  
Based Peak Detection Routine.  
Default values for different types of data are:  
Mariner data5,000, which is optimized for masses  
below 3,000 m/z. Decrease this value if you are  
analyzing proteins.  
Voyager linear data2,000, which is optimized for  
masses below 20,000 m/z. Decrease this value if you  
are analyzing higher masses or want to label average  
isotope masses.  
3
Voyager reflector data10,000, which is optimized for  
masses below 20,000 m/z. Decrease this value if you  
are analyzing higher masses or want to label average  
isotope masses.  
If you set a resolution value of:  
1,000 or greaterThe software uses the setting until it  
reaches the mass at which isotopic resolution is no  
longer possible, then switches to a resolution of 1,000,  
the resolution that corresponds to isotopic clusters.  
NOTE: The Mass Resolution you set here is also used by the  
Elemental Composition Calculator, the Elemental Targeting  
Application, and the Default Smoothing function. For more  
information, see Section 6.1, Using the Elemental  
Composition Calculator, Section 6.6, Using the Elemental  
Targeting Application, and Section 5.7, Noise  
Filtering/Smoothing.  
Continued  
3-24  
Applied Biosystems  
 
Peak Detection  
Table 3-2 Basic Settings Tab Parameters (Spectrum Data Only) (Continued)  
Parameter  
Description  
Trace Settings  
Use same settings  
for all traces in view  
Applies settings to all traces in the active window.  
NOTE: In previous versions of Data Explorer software,  
peak detection allowed you to specify Peak Width. The  
software now automatically uses a minimum peak width that  
is equal to the Filter Width and a maximum peak width of  
10,000 data points.  
3
Data ExplorerSoftware Users Guide  
3-25  
   
Chapter 3 Peak Detection and Labeling  
Peak Processing Table 3-3 describes the parameters in the Peak Processing  
tab of the Spectrum Peak Detection Setup dialog box (see  
Figure 3-5 on page 3-16).  
parameters  
(spectrum data  
Default peak detection values are listed in Section 3.7, Default  
Peak Detection Settings.  
only)  
Table 3-3 Peak Processing Parameters (Spectrum Data Only)  
Parameter  
Description  
Integration Baseline Settings  
NOTE: You can set peak labels to display peak start, peak end, and baseline  
marks. See Section 3.5.2, Setting Chromatogram and Spectrum Peak Labels.  
Valley-to-Baseline  
Valley-to-Valley  
Draws a vertical line from all valleys to a horizontal baseline.  
The level of the horizontal baseline is determined using the  
minimum peak valley point (left or right) for each peak. See  
Figure 3-25 on page 3-70.  
3
Forces a baseline through all valley points. See Figure 3-25  
on page 3-70.  
Spectrum Parameters  
Centroid %  
Specifies the percentage of peak height used to determine  
the location of the centroid mass. For example, with a  
Centroid % of 10, the software examines the top ten percent  
of signal.  
Enable Gaussian  
Fitting  
Increases the accuracy of the centroid calculation for  
spectral peaks with limited data points (fewer than 6 data  
points across the peak), in particular, Mariner data below  
m/z 400.  
Continued  
3-26  
Applied Biosystems  
                       
Peak Detection  
Table 3-3 Peak Processing Parameters (Spectrum Data Only) (Continued)  
Parameter  
Description  
Charge State Determination (Spectrum only)  
Maximum Charge  
State  
Determines the peak spacing evaluated for the presence of  
isotope peaks. The expected peak spacing is determined by  
the Max Charge State plus or minus a tolerance value. The  
tolerance is calculated as (proton mass/charge state) × 15%.  
For example, with a Max Charge State of 3, the software  
checks for peaks that are 0.33 m/z apart, with a fixed  
tolerance of 0.05 m/z, where proton mass equals  
1.007276456. Maximum Charge State is 6 for Mariner data  
and 1 for Voyager data.  
Maximum Isotopes # Specifies the maximum number of peaks included in an  
isotope cluster.  
NOTE: Setting this value too low can result in peaks not  
being included in the appropriate isotope cluster. For more  
information, see Max Isotope # set too lowon page 3-35.  
3
Minimum and  
Specifies the intensity range that a peak must fall within  
Maximum Intensity% (relative to the previously evaluated higher mass peak) to be  
included in the current isotope cluster.  
Data ExplorerSoftware Users Guide  
3-27  
         
Chapter 3 Peak Detection and Labeling  
Advanced Table 3-4 describes the parameters in the Advanced Settings  
tab of the Spectrum Peak Detection Setup dialog box (see  
Figure 3-6 on page 3-17).  
Settings  
(spectrum data  
only)  
Default peak detection values are listed in Section 3.7, Default  
Peak Detection Settings.  
NOTE: If you select Use Resolution Dependent Settings in  
the Basic Settings tab, Basic Settings override Advanced  
Settings. The Advanced Settings tab is accessible, but all  
parameters on the tab are dimmed. To make Advanced  
Settings available for editing, select Use Advanced  
Settings on the Basic Settings tab.  
Table 3-4 Advanced Settings (Spectrum Data Only)  
Parameter  
Peak Detection Settings  
Detection Ranges  
Description  
3
Specifies the region of the trace (defined by x-axis lower and  
upper boundaries) to which the settings apply.  
If you select Use Resolution Dependent Settings in the Basic  
Settings tab (described on page 3-23), the software  
automatically divides the trace into detection ranges. If you  
select Use Advanced Settings, the detection ranges  
calculated by the software are maintained, and you can  
modify the ranges and settings, or combine all ranges in the  
list into one range, as described below.  
Continued  
3-28  
Applied Biosystems  
                     
Peak Detection  
Table 3-4 Advanced Settings (Spectrum Data Only) (Continued)  
Parameter  
Description  
Detection Ranges  
(continued)  
You can set multiple, non-contiguous ranges and define  
parameters for each range independently.  
You select a range in the Detection Ranges list box by  
single-clicking the range number.  
To add a detection range, do one of the following:  
Select an existing range, then click  
. This creates a  
new range with boundaries ranging from the end of the  
existing range to the end of the trace. If the existing  
range ends at the end of the trace, the region of the  
existing range is split in half between the existing range  
and the new range.  
Double-click an existing range to manually enter lower  
and upper boundaries.  
3
Select a range in the dialog box, then click-drag the  
X data cursors labels in the trace to set the lower and  
upper boundaries.  
To delete a range, select the range, then click  
.
To combine all ranges in the list into one range, click  
.
The peak detection settings displayed in the dialog box  
correspond to the selected range. To view peak detection  
settings for another range, select the range of interest.  
Continued  
Data ExplorerSoftware Users Guide  
3-29  
     
Chapter 3 Peak Detection and Labeling  
Table 3-4 Advanced Settings (Spectrum Data Only) (Continued)  
Parameter  
Description  
Active Range Thresholds  
NOTE: These settings apply to the Detection Range selected, and override the  
Global Thresholds specified on the Basic Settings tab (described on page 3-19).  
%BP Intensity  
See %Base Peak Intensityon page 3-20.  
%Max Peak Area  
Minimum Intensity  
See %Max Peak Areaon page 3-20.  
Specifies the absolute peak intensity below which peaks are  
not detected. Calculated relative to zero.  
NOTE: Maximum possible intensity for Mariner data is  
11,000 counts (spectrum that has not been accumulated or  
summed), and for Voyager data is 66,000 counts.  
NOTE: This parameter was previously named Absolute  
Threshold.  
3
Minimum Area  
Specifies the peak area below which peaks are not detected.  
Calculated relative to peak valleys.  
Hint: Display the peak list to determine the appropriate area  
value to enter.  
Noise Threshold  
Used to determine peak boundaries. Click the field, then  
right-click-drag over a segment of the trace to automatically  
set.  
For more information, see Process that Occurs During Peak  
Detection, Centroiding, and Integrationon page 3-67.  
Continued  
3-30  
Applied Biosystems  
                   
Peak Detection  
Table 3-4 Advanced Settings (Spectrum Data Only) (Continued)  
Parameter  
Filter Settings  
Width  
Description  
Number of data points used in smoothing for peak detection  
before integration. This value is automatically calculated by  
the software if you select Use Resolution Dependent Settings  
on the Basic Settings tab. For more information, see  
Section 3.1.2, The Resolution-Based Peak Detection  
Routine.  
NOTE: If you set Filter Width too high, narrow peaks may not  
be detected.  
to the number of points across the peak. To determine the  
number of points across a peak, you can change the trace  
display from Line to Vertical Bars. Each vertical bar  
represents one data point. For more information, see  
Section 1.4, Customizing the Data Explorer Window.  
3
Increment  
Number of data points the filter moves across. Use 1 for all  
applications.  
This value is automatically set to 1 if you enable Use  
Resolution Dependent Settings on the Basic Settings tab.  
For more information, see Section 3.1.2, The Resolution-  
Based Peak Detection Routine.  
Data ExplorerSoftware Users Guide  
3-31  
         
Chapter 3 Peak Detection and Labeling  
3.2.5 Charge State  
NOTE: Isotope-resolved peaks in Voyager data are  
typically singly charged. See Section 3.7, Default Peak  
Detection Settings, for recommended settings.  
This section includes:  
Charge state determination  
Charge state parameter examples  
Charge state determination troubleshooting  
Isotopes For information on isotopes, see Appendix B, Overview of  
Isotopes.  
Charge state Charge state is determined by evaluating the relative isotope  
peak spacing. The expected peak spacing is determined by  
determination  
3
the Max Charge State, plus or minus a tolerance value. The  
tolerance is calculated as (proton mass/charge state) × 15%,  
where proton mass equals 1.007276456.  
During charge state determination, the software starts with the  
most intense peak and the peak spacing determined by the  
Max Charge state, then evaluates all detected peaks. The  
software iteratively decreases the Max Charge State and  
repeats the process for all peaks not already identified as  
members of an isotope cluster.  
Charge state To accurately determine charge state, Charge State  
parameters must be set appropriately. The following sections  
illustrate:  
parameter  
examples  
Max Charge State and Max Isotope # set correctly  
Max Charge State set too low  
Max Isotope # set too low  
Effect of Minimum Intensity  
3-32  
Applied Biosystems  
         
Peak Detection  
Max Charge State When the Max Charge State and Max Isotope # are set  
correctly (in the example shown in Figure 3-7, both are set  
to 4), the neurotensin 558 m/z isotope cluster contains four  
peaks at charge state 3. Peaks are 0.33 m/z apart  
(Figure 3-7).  
and Max Isotope #  
set correctly  
Figure 3-7 Max Charge State and Max Isotope # Set  
Correctly for Neurotensin  
Max Charge State When the Max Charge State is set too low, the charge state for  
all peaks may not be correctly determined. In this example  
set too low  
with the Max Charge State set to 2, only the first and fourth  
peaks are labeled with charge states (Figure 3-8 on  
page 3-34). This is because the software:  
Checks for peaks at 0.5 m/z from each peak (charge  
state 2 evaluation).  
Finds no peaks and checks for peaks at 1 m/z from each  
peak (charge state 1 evaluation).  
Finds one peak pair at 558.3 m/z and 559.3 m/z and  
labels these peaks as charge state 1.  
Does not find a match for peaks at 558.6 m/z or 558.9  
m/z, so these peaks are considered to have no calculated  
charge, and are not labeled with a charge state  
(Figure 3-8 on page 3-34).  
Data ExplorerSoftware Users Guide  
3-33  
     
Chapter 3 Peak Detection and Labeling  
.
Figure 3-8 Max Charge State Set Too Low and Max  
Isotope # Set Correctly  
The grouping of the first and fourth peaks into an isotope  
cluster is apparent when you turn on monoisotopic peak list  
filtering (Figure 3-9). The software labels the first peak, but  
removes the mass and charge state labels from the 559.3 m/z  
peak, indicating that it is part of the 558.3 m/z isotope cluster.  
3
Note that turning on monoisotopic peak list filtering does not  
affect the mass labels on the 558.6 m/z or 558.9 m/z peaks,  
because these peaks have no calculated charge and are not  
part of the determined isotope cluster.  
.
Figure 3-9 Max Charge State Set Too Low,  
Monoisotopic Peak List Filtering Turned On  
3-34  
Applied Biosystems  
   
Peak Detection  
Max Isotope # If the Max Isotope # is set too low, the software incorrectly  
groups peaks into different isotope clusters. In this example  
with the Max Isotope # set to 2 and Max Charge State set to 4,  
the software labels all peaks as charge state 3 (Figure 3-10).  
set too low  
Figure 3-10 Max Isotope # Set Too Low, Max Charge  
State Set Correctly  
3
The effect of setting Max Isotope # too low is apparent when  
you turn on Monoisotopic peak list filtering. Figure 3-11 shows  
that the software has grouped the peaks into two clusters and  
has identified two monoisotopic peaks.  
Figure 3-11 Max Isotope # Set Too Low, Monoisotopic  
Peak List Filtering Turned On  
Data ExplorerSoftware Users Guide  
3-35  
       
Chapter 3 Peak Detection and Labeling  
Effect of Minimum The Minimum Intensity setting on the Advanced Settings tab  
can also determine how charge states are determined for a  
Intensity  
peak, because it determines if the software can find a match  
for a peak. Figure 3-12 illustrates two peaks correctly labeled  
with charge state 1.  
Figure 3-12 Minimum Intensity Set Correctly  
If you increase the Minimum Intensity setting to 85 in the  
example shown in Figure 3-12, the software cannot determine  
a charge state for the first peak (Figure 3-13). This is because:  
3
The increased threshold suppresses the detection of the  
236.6 m/z peak.  
When the software evaluates the peak at 235.6 m/z, it  
does not find a peak at the appropriate peak spacing for  
charge state 1, and therefore determines that the peak  
has no calculated charge.  
Figure 3-13 Minimum Intensity Set Incorrectly  
Charge state See Table 9-12, Charge State and Isotope Determination  
TroubleshootingMariner Only,on page 9-18.  
determination  
troubleshooting  
3-36  
Applied Biosystems  
     
Peak List  
3.3 Peak List  
This section describes:  
Displaying the peak list  
Inserting peaks in the peak list  
Saving the peak list  
Sorting, filtering, and printing the peak list  
3.3.1 Displaying the Peak List  
After peak detection, centroiding, and integration, the software  
creates a peak list for the chromatogram (Mariner data only)  
and each spectrum in the data file.  
Displaying To display the peak list, select Output Window from the View  
menu, then click the Chro Peak List or Spec Peak List tab at  
the bottom of the window (Figure 3-14).  
3
Click the Peak List tab  
at the bottom of  
the Output window  
Figure 3-14 Peak List in Output Window  
Data ExplorerSoftware Users Guide  
3-37  
             
Chapter 3 Peak Detection and Labeling  
Contents The peak lists contain the following information.  
of peak list  
Chromatogram Each entry represents one chromatographic peak in the trace  
and includes:  
peak list  
Index, a sequential number assigned to each entry in  
the peak list  
Spectrum # or retention time (minutes), determined by  
the command selected from the Display menu  
Lower bound (spectrum # or minutes)  
Upper bound (spectrum # or minutes)  
Peak height (calculated relative to zero)  
Peak area (calculated relative to local baseline)  
Spectrum peak list Each entry represents one spectral peak in the trace and  
includes:  
3
Index, a sequential number assigned to each entry in  
the peak list  
Centroid or Apex mass (reflects the Mass Type  
selected in the Peak Label dialog box)  
Lower bound (m/z)  
Upper bound (m/z)  
Charge (z)  
Peak height  
Relative intensity (calculated relative to zero, not to  
baseline)  
Peak area (calculated relative to local baseline)  
3-38  
Applied Biosystems  
       
Peak List  
3.3.2 Inserting Peaks in the Peak List  
Description If chromatogram or spectrum peaks are not detected and  
labeled by the selected detection parameters, you can  
manually detect and label peaks by inserting peaks in the  
peak list.  
Procedure To insert peaks:  
1. Display the chromatogram or spectrum trace of  
interest.  
2. From the Peaks menu, select Insert Peaks.  
The Insert Peaks dialog box is displayed (Figure 3-15).  
3
Figure 3-15 Insert Peaks Dialog Box  
3. To enter the Left Edge and Right Edge of the peak to  
insert, right-click-drag over the region of the trace to  
calculate, or type in masses.  
4. Click Calculate.  
The peak is inserted in the peak list displayed in the  
Peak list tab in the Output window, and centroid  
information is listed in the Result tab. For more  
information, see Output windowon page 1-15.  
Data ExplorerSoftware Users Guide  
3-39  
         
Chapter 3 Peak Detection and Labeling  
Inserted peaks are:  
Removed from the list when you close the data  
file, reprocess the data, or set peak detection  
parameters so that the inserted peak is no longer  
detected.  
Assigned a charge state of 0 to indicate the  
charge state is unknown.  
CAUTION  
A zero value in the Spec Peak list does not indicate a  
charge state of zero. It indicates that the software  
could not determine the charge state.  
3.3.3 Saving the Peak List  
3
Saving as a You can save the contents of the chromatogram and spectrum  
stand-alone  
peak lists as stand-alone peak list files (.PKT).  
.PKT file  
Stand-alone peak list files can be used in other applications  
such as Microsoft Notepad Editor or Microsoft Excel.  
To save a peak list as a stand-alone .PKT file for use in other  
applications:  
1. Click the trace of interest.  
2. Display the peak list of interest in the Output window by  
clicking the Chro Peak List or the Spec Peak List tab.  
3. Right-click the Output window, then select Save As.  
The Save Peak Table As dialog box is displayed.  
4. Select a directory and type a file name.  
5. Click Save.  
The software automatically assigns a .PKT extension  
and saves the peak list as a tab-delimited text file.  
3-40  
Applied Biosystems  
             
Peak List  
NOTE: Inserted peaks are included when you save a  
.PKT file.  
NOTE: Peak list headings are not included when you save  
a .PKT file. If you require headings, copy the peak list  
directly to Excel instead of saving it as a .PKT file and then  
importing it. Right-click the peak list to copy.  
Importing and You can import peak list values into Microsoft Excel to create a  
spreadsheet.  
saving in Excel  
To import and save the peak list in Microsoft Excel:  
1. In Excel, select Open from the File menu.  
The Open dialog box is displayed.  
2. Select All Files (*.*) from the Files of Type drop-down list  
at the bottom of the dialog box. Select the directory and  
file to import, then click Open.  
3
The Text Import wizard is displayed.  
3. Follow the prompts from the Text Import wizard, accepting  
the default settings provided.  
The peak list is converted to an Excel spreadsheet.  
4. Select Save As from the File menu.  
The Save As dialog box is displayed.  
5. Select Microsoft Excel Workbook (*.xls) in the Save as  
Type drop-down list at the bottom of the dialog box.  
6. Type a file name or modify the default name (delete  
quotation marks, if present, and the .PKT extension).  
7. Click Save.  
The file is saved with an .XLS extension.  
For more information on using Excel, refer to your  
Microsoft Excel Users Guide.  
Data ExplorerSoftware Users Guide  
3-41  
       
Chapter 3 Peak Detection and Labeling  
and Printing the Peak List  
Sorting  
the peak list  
The peak list is displayed in order by index number. You can  
sort the list by any field by clicking the column header buttons  
(Figure 3-16).  
Click column header buttons to sort by different fields  
3
Figure 3-16 Sorting the Peak List  
Filtering Peak list filtering allows you to display only desired peaks in  
the spectrum peak list. Only peaks that are included in the  
peak list are labeled on the trace.  
the spectrum  
peak list  
To filter the peak list:  
1. From the Peaks menu, select Filter Peak List.  
Hint: You can also display the Mass Peak List Filter  
dialog box by right-clicking the Spec Peak List, then  
selecting Peak Filter.  
3-42  
Applied Biosystems  
               
Peak List  
2. Select Enable Peak List Filter, then select:  
Filter Type  
Description  
Monoisotopic Labels the peak of the lowest detected mass in an isotope  
envelope. Before applying filtering, set the values in Peak  
Processing detection parameters to accurately detect and label all  
peaks in the isotope envelope with the correct charge state. If the  
parameters are not correctly set to yield correct charge states for  
all peaks, the monoisotopic peak may not be correctly labeled.  
See Section 3.2.5, Charge State Determination and Examples, for  
more information on setting parameters.  
Monoisotopic filtering is useful on isotope clusters where the  
approximate elemental composition and isotopic ratios are not  
known, for example, in PSD analysis (where the precursor ion  
filtering enhances the precursor ion isotope intensity relative to the  
rest of the cluster), or in a spectrum that contains molecules with  
very different elemental compositions.  
3
For most applications, especially peptide analysis, the deisotoping  
function yields more accurate results than monoisotopic peak  
filtering. See Section 3.4, Deisotoping a Spectrum.  
Charge State  
Labels the peaks with the selected charge state (z). If you select a  
charge state of 0, the software displays mass labels only on peaks  
for which the software could not determine a charge state.  
CAUTION: A zero value in the Spec Peak list does not indicate a  
charge state of zero. It indicates that the software could not  
determine the charge state.  
Data ExplorerSoftware Users Guide  
3-43  
               
Chapter 3 Peak Detection and Labeling  
Printing To print the peak list:  
the peak list  
1. Display the peak list as you want it printed.  
2. Right-click the peak list, then select Print.  
Deleting items To delete an item from the peak list, select the item, right-click  
the spectrum peak list, then select Delete Peak. The entry is  
removed from the list, and the peak label is removed from the  
trace.  
from the peak list  
To display the deleted item in the peak list again, select Peak  
Detection from the Peaks menu, then click OK.  
3
3-44  
Applied Biosystems  
   
Deisotoping a Spectrum  
3.4 Deisotoping a Spectrum  
This section includes:  
Description  
During peak deisotoping  
When to use  
Requirements  
Using the Deisotope function  
Troubleshooting  
Example  
Returning to the original spectrum  
Description The Deisotope function reduces a spectrum to a centroided  
plot by deconvoluting the monoisotopic peaks from the current  
peak list.  
Peak deisotoping is an advanced peak filtering method that  
can determine the relative abundance of multiple components  
with overlapping isotope distributions. The deisotoping  
algorithm uses the elemental composition that you specify to  
improve the determination of the monoisotopic mass by  
considering the centroid masses of all peaks in the isotopic  
envelope.  
3
During peak For each detected peak in a spectrum, the software inspects  
the peak list for the higher theoretical masses and areas  
associated with additional expected peaks in a theoretical  
isotopic cluster, relative to the peak in question.  
deisotoping  
Data ExplorerSoftware Users Guide  
3-45  
         
Chapter 3 Peak Detection and Labeling  
If the expected higher theoretical peak masses and areas are  
present in the peak list:  
The peak in question is considered to be a monoisotopic  
peak.  
The intensities of the higher mass peaks that correspond  
to the expected isotope ratios are combined with the  
intensity of the peak in question (additional intensity that  
may be related to a contaminant or an overlapping  
isotope envelope is not combined, and will be evaluated  
in the next iteration).  
The peak in question is represented in the trace as a  
centroid bar with increased amplitude. The total intensity  
of the centroid bar represents the total area of each fitted  
cluster.  
If the expected higher theoretical peak masses and areas are  
not present in the peak list, the peak in question is  
represented in the trace as a centroid bar with its original  
amplitude.  
3
Figure 3-17 illustrates how peaks that are and peaks that are  
trace.  
When to use For most applications, particularly peptide analysis, the  
deisotoping function yields more useful results than  
monoisotopic peak filtering (described in Section 3.3.4,  
Sorting, Filtering, and Printing the Peak List) because the  
deisotoping function:  
Can successfully identify the monoisotopic peaks in  
overlapping clusters.  
Does not consider noise peaks that exhibit the mass, but  
not the area of an expected isotope peak.  
Amplifies the intensity of monoisotopic masses at high  
m/z (due to contribution from other peaks in the cluster),  
which allows the peak detection thresholds to eliminate  
chemical noise without eliminating the high m/z peaks of  
interest, enables improved peak matching in calibration,  
and provides better results in database searching (which  
relies on monoisotopic masses).  
3-46  
Applied Biosystems  
Deisotoping a Spectrum  
Peaks in original trace  
Deisotoped trace  
Increased amplitude of first  
peak indicates it is a  
monoisotopic peak  
If first three peaks are part of  
the same isotope cluster  
Peak that is not  
part of cluster  
Contribution to expected  
isotope ratio  
by contaminant  
Increased amplitude of first  
peak indicates it is a  
monoisotopic peak  
If first three peaks are part of  
the same isotope cluster, but  
contaminant is also present  
Contaminant  
Peak that is not  
part of cluster  
Increased amplitude of peaks  
indicate they are  
monoisotopic peaks  
If two isotope clusters  
are present  
3
Figure 3-17 Interpreting a Deisotoped Trace  
Requirements The Deisotope function requires a singly charged spectrum. If  
the spectrum you are examining includes multiply charged  
peaks, use the Single-Charge Conversion function on the  
Process menu before using the Deisotope function. For  
information, see Section 5.10, Converting to a Singly Charged  
Spectrum (Mariner Data Only).  
CAUTION  
If you use the Deisotope function on multiply charged  
peaks, invalid results are reported.  
Data ExplorerSoftware Users Guide  
3-47  
   
Chapter 3 Peak Detection and Labeling  
Using the To use the Deisotope function:  
Deisotope  
function  
1. Display the spectrum trace of interest.  
2. Make sure peak detection thresholds are set low enough  
to detect the monoisotopic peak before deisotoping. If the  
detection thresholds are not set low enough, adjust them.  
Parameters.  
If you are analyzing digest data, set %Max Peak Area  
interest are detected. For more information, see  
Section 7.2.3, Detecting Peaks from Complex Digests.  
3. If multiply charged peaks are present, convert to a singly  
charged spectrum. See Section 5.10, Converting to a  
Singly Charged Spectrum (Mariner Data Only).  
4. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
3
5. From the Peaks menu, select Peak Deisotoping.  
The Deisotoping dialog box (Figure 3-18) is displayed.  
Figure 3-18 Deisotoping Dialog Box  
6. In the Adduct text box, type the adduct that is the  
charge-carrying species in the spectrum you are  
examining.  
3-48  
Applied Biosystems  
   
Deisotoping a Spectrum  
CAUTION  
If you enter an invalid value in the Adduct field, for  
example, numbers, the spectrum is still converted to a  
deisotoped spectrum, and the peak height is  
proportional to the original peak area.  
7. In the Formula text box, type the generic formula (using  
any elements in the Periodic table) that approximates the  
isotopic pattern for the compound class you are  
examining. Generic formulas include:  
PeptidesC6H5NO  
DNAC38H49N15O24P4  
RNAC38H47N15O28P4  
CarbohydrateC6H12O6  
8. Click OK.  
The deisotoped spectrum is displayed with a DI trace  
label. The mass of the vertical bar corresponds to the  
mass of the monoisotopic peak. The height of the  
vertical bar is proportional to the peak areas of all  
peaks in the isotope cluster.  
3
Troubleshooting If you see several small peaks around a very large peak after  
deisotoping, the software has successfully identified a  
monoisotopic peak, but the isotope cluster does not exactly  
match the theoretical elemental composition that you  
specified. This may be caused by the following:  
The signal-to-noise ratio of the spectrum is poor and the  
calculated peak areas are not accurate.  
The specified elemental composition is not correct for the  
compound. For example, if the peptide contains sulfur,  
and sulfur was not in the specified formula, a small peak  
may occur 2 Da higher than the expected monoisotopic  
mass.  
There are two overlapping isotope clusters but the  
intensity of one cluster is stronger than the intensity of the  
other.  
Increase the peak detection threshold after deisotoping to  
remove the residual peaks from the peak list.  
Data ExplorerSoftware Users Guide  
3-49  
   
Chapter 3 Peak Detection and Labeling  
Example Figure 3-19 and Figure 3-20 illustrate the effects of  
deisotoping.  
Before deisotoping (Figure 3-19), the spectrum includes an  
isotope pattern with four detected peaks.  
Figure 3-19 Spectrum Before Deisotoping  
After deisotoping (Figure 3-20), the trace includes two  
labeled centroid bars that represent monoisotopic masses,  
indicating that the original trace represents two isotopic  
envelopes. The intensity of a centroid bar is proportional to  
the area of all peaks in the isotope cluster.  
3
Figure 3-20 Spectrum After Deisotoping  
3-50  
Applied Biosystems  
     
Deisotoping a Spectrum  
Returning to the To return to the original spectrum:  
original spectrum  
If the original spectrum was an unprocessed spectrum,  
select Spectrum Number from the Display menu. The  
number of the original spectrum is displayed in the Select  
Spectrum dialog box. Click OK.  
If the original trace was a processed spectrum, select  
Processing History from the Display menu, then select  
the original trace.  
3
Data ExplorerSoftware Users Guide  
3-51  
Chapter 3 Peak Detection and Labeling  
3.5 Peak Labeling  
This section includes:  
Charge state labels  
Setting chromatogram and spectrum peak labels  
Setting custom peak labels  
list (displayed in the Output window). Thresholds for peak  
labeling are set independent of peak detection.  
peak labeling  
Peak detection parameters, described in Section 3.6,  
Process that Occurs During Peak Detection, Centroiding,  
and Integration  
Charge state determination, described in Section 3.2.5,  
Charge State Determination and Examples  
Peak list filtering, described in Filtering the spectrum  
3
peak liston page 3-42  
NOTE: If peak labeling is disabled, no labels are displayed,  
even for peaks in the peak list. If you delete a peak from the  
peak list, it is not labeled.  
Labeling peaks If peak detection settings do not detect and label desired  
peaks, you can also manually label peaks by inserting them  
into the peak list. See Section 3.3.2, Inserting Peaks in the  
Peak List.  
manually  
3-52  
Applied Biosystems  
                 
Peak Labeling  
3.5.1 Charge State Labels  
Charge state A unique feature of the Data Explorer software is the ability to  
label multiply charged isotope peaks with their charge state.  
labels  
In general, the observed spacing between isotopes is  
determined by the charge state of the detected ion. Spacing  
between isotope peaks is narrower at higher charge states.  
For more information on isotopes, see Appendix B, Overview  
of Isotopes.  
To determine charge state, the software uses the following  
Max Isotope #  
Min and Max Intensity  
For a description of these parameters, see Advanced  
Settings (spectrum data only)on page 3-28 and Charge  
NOTE: If Charge State Determination parameters are set  
incorrectly, the charge state is determined incorrectly. For  
more information, see Charge state parameter examples”  
on page 3-32 and Charge State and Isotope Determination  
TroubleshootingMariner Onlyon page 9-18.  
Requirements To label isotope peaks with charge states, the following  
must be true:  
Mass must be within the range in which the system can  
resolve the isotope peaks:  
<4,000 Da on a Mariner system  
<5,000 Da on a Voyager-DE PRO system  
<6,000 Da on a Voyager-DE STR system  
Data ExplorerSoftware Users Guide  
3-53  
         
Chapter 3 Peak Detection and Labeling  
Isotopes must be resolved  
Peak detection parameters must be set to detect all of the  
peaks in the isotope cluster  
Charge determination parameters must be set  
appropriately  
Charge state labels must be enabled  
3.5.2 Setting Chromatogram  
and Spectrum Peak Labels  
This section includes:  
Customizing colors, font, or size  
Setting chromatogram labels  
Deleting labels  
Labels not displayed  
Charge state not displayed  
3
Customizing To set colors, font, and size for peak labels, see Section 1.5,  
Setting Graphic Options.  
colors, font, and  
size  
Setting To label chromatogram peaks:  
chromatogram  
labels  
1. Click the Chromatogram window to activate it.  
2. From the Peaks menu, select Peak Label.  
The Chromatogram Peak Label dialog box is displayed  
(Figure 3-21).  
3-54  
Applied Biosystems  
             
Peak Labeling  
Figure 3-21 Chromatogram Peak Label Dialog Box  
3
3. Select Enable Labeling.  
4. Set the number of decimal points to be displayed.  
5. Select Label Attributes:  
OverlappingAllows labels to be displayed when  
peaks are close together.  
Peak boundsDisplays peak start, peak end,  
and baseline.  
OrientationSpecifies Horizontal, 45-degree, or  
Vertical labels.  
6. Select the label content to display, (for example,  
Spectrum #, Time, Vial Number).  
7. To create custom labels, select User Labels, then click  
User Label Setup. See Section 3.5.3, Setting Custom  
Peak Labels.  
Data ExplorerSoftware Users Guide  
3-55  
                 
Chapter 3 Peak Detection and Labeling  
8. Click OK.  
The trace is displayed. The detected peaks that meet  
the peak labeling criteria are labeled.  
Setting To label spectrum peaks:  
spectrum labels  
1. Click the Spectrum window to activate it.  
2. From the Peaks menu, select Peak Label.  
The Spectrum Peak Label dialog box is displayed  
(Figure 3-22).  
3
Figure 3-22 Spectrum Peak Label Dialog Box  
3. Select Enable Labeling to label peaks in m/z format.  
4. Set the number of decimal points to be displayed.  
3-56  
Applied Biosystems  
             
Peak Labeling  
5. Select the Mass Type (peak apex or peak centroid).  
6. Select the Peak Mass Label Type:  
Label  
Mass  
Description  
Labels with Apex or Centroid mass.  
NOTE: If you create a custom user label for a mass, the user label is  
displayed instead of the mass. For more information, see  
Section 3.5.3, Setting Custom Peak Labels.  
Mass  
Shifts all peak labels by the value entered (either positive or  
negative) and can be used to calculate mass differences from a  
reference peak.  
difference  
from the  
selected  
peak  
Right-click-drag across the reference peak to enter the negative of  
the value of the peak in the Mass difference text box, then click OK.  
NOTE: This The reference peak is labeled with zero, and all the other labels are  
label was  
previously  
called Mass  
Offset.  
plus or minus their mass difference from the reference peak.  
3
This label is useful to display neutral losses and adducts on a given  
peak.  
Mass  
Labels peaks with a mass relative to the adjacent labeled peak of  
lower m/z.  
difference  
between  
adjacent  
peaks  
Data ExplorerSoftware Users Guide  
3-57  
       
Chapter 3 Peak Detection and Labeling  
7. Select label attributes:  
OverlappingAllows labels to be displayed when  
peaks are close together.  
Peak boundsDisplays peak start, peak end,  
and baseline.  
OrientationSpecifies Horizontal, 45-degree, or  
Vertical labels.  
8. Select label content to be displayed in addition to m/z:  
AreaIntegrated area of peak, displayed with an  
Alabel.  
Charge StateCharge state of peak, displayed  
with a zlabel, where the zrepresents a positive  
or negative charge.  
3
NOTE: A zis displayed for positive or negative ions.  
To determine the actual charge of the ions, display the  
Instrument Setting tab in the Data Explorer Output  
window. See Output windowon page 1-15 for  
information. Check the listed polarity. For spectra  
acquired in positive ion mode, positive ions are  
produced. For spectra acquired in negative ion mode,  
9. To create custom labels, select User Labels, then click  
User Label Setup. See Section 3.5.3, Setting Custom  
Peak Labels.  
10. Click OK.  
The trace is displayed.  
3-58  
Applied Biosystems  
             
Peak Labeling  
that meet the peak filtering criteria are labeled  
according to the peak label settings. Otherwise, all  
detected peaks are labeled according to the peak label  
settings. See Section 3.2.4, Peak Detection  
Parameter Descriptions, and Filtering the spectrum  
peak liston page 3-42.  
Deleting labels To delete a peak label from the trace, click the Spec Peak List  
tab in the Output window, select the item, right-click the  
spectrum peak list, then select Delete Peak.  
Labels not If a label is not displayed or is displayed incorrectly, possible  
causes are:  
displayed  
The label is too long, peaks are too close together, or a  
peak is very close to the right axis. If there is not enough  
room for the label to be displayed, the label is  
suppressed. Zooming the region of interest expands the  
trace and may allow the labels to display.  
3
To display labels when peaks are close together, select  
Allow overlapping peak labels in the Peak Label dialog  
box.  
Peak filtering is enabled, labeling only peaks that meet  
the peak list filtering criteria.  
Filter Width is set too high to detect the peak.  
Data ExplorerSoftware Users Guide  
3-59  
   
Chapter 3 Peak Detection and Labeling  
Charge state If no charge is displayed, there are a few possible causes:  
not displayed  
Peaks are more than 1 Da apart.  
Filter width is set too high to detect other isotope peaks.  
determination is set lower than the charge state of the  
peak.  
Charge state determination parameters are set such that  
peaks are determined to have no charge. See  
Section 3.2.5, Charge State Determination and  
Examples.  
Peaks are not from the same isotope species.  
NOTE: If you set %Base Peak Intensity or %Max Peak  
Area too low (described on page 3-20), noise peaks  
can be detected, and charge states can be incorrectly  
labeled.  
3
3-60  
Applied Biosystems  
   
Peak Labeling  
3.5.3 Setting Custom Peak Labels  
This section includes:  
Description  
Customizing colors, font, and size  
Creating custom peak labels  
Applying user labels from .LBC or LBS files  
Displaying user labels  
User labels not displayed  
Description A custom peak label displays the label name you enter. If you  
specify a custom label for a spectrum, it is displayed instead of  
the mass value.  
Customizing To set colors, font, and size for custom labels or peak labels,  
see Section 1.5, Setting Graphic Options.  
colors, font, and  
size  
Creating custom To create custom peak labels:  
3
peak labels  
1. Click the Chromatogram or Spectrum window to  
activate it.  
2. From the Peaks menu, select Peak Label.  
The Peak Label dialog box is displayed (see  
Figure 3-22 on page 3-56).  
3. Click User Label Setup.  
The User Label Setup dialog box appears  
(Figure 3-23).  
NOTE: The Chromatogram User Label Setup dialog  
box displays X Value and X Tolerance instead of Peak  
Mass and Mass Error.  
Data ExplorerSoftware Users Guide  
3-61  
         
Chapter 3 Peak Detection and Labeling  
Column  
header  
buttons  
Figure 3-23 User Label Setup Dialog Box  
3
4. Select the Label Type (spectra only):  
MassLabels with Apex or Centroid mass.  
Mass difference from adjacent peaksApplies  
the labels to peaks that have the specified mass  
difference relative to the adjacent labeled peak of  
lower m/z.  
5. Select Match Charge State (spectra only) if you want the  
charge state of a peak evaluated before applying the user  
label. When this function is enabled, a peak must occur  
within the specified Mass Tolerance (described below),  
and have the specified charge state before the user label  
is applied.  
Hint: The Match Charge State function allows you to  
screen out peaks that are within the specified Mass  
Tolerance, but are not the charged species you are  
interested in.  
3-62  
Applied Biosystems  
       
Peak Labeling  
6. To manually enter label settings, click  
.
The User Label Entry dialog box is displayed  
(Figure 3-24).  
Figure 3-24 User Label Entry Dialog Box (Spectrum)  
7. Set the following parameters:  
3
Spectrum Parameter  
Description/Specifies  
Text of the label to display.  
Label  
Peak Mass (if Label Type selected  
is Mass)  
Mass of the peak to which the label applies.  
Mass Difference (if Label Type  
selected is Difference)  
Difference in mass that must exist between  
peaks to apply user labels.  
Mass Tolerance  
Mass tolerance that the peak must occur  
within to apply user labels.  
NOTE: If a peak meets the criteria for more  
than one user label, multiple user labels are  
applied to the peak and are separated by  
commas.  
Charge State  
Charge state that the peak must have in  
order to apply user labels.  
Data ExplorerSoftware Users Guide  
3-63  
 
Chapter 3 Peak Detection and Labeling  
Chromatogram Parameter  
Description  
Label  
Text of the label to display.  
X Value  
Retention time or Spectrum Number of the  
peak to label. The default units for X  
Tolerance correspond to the units of the  
x-axis.  
X Tolerance  
Retention time or Spectrum Number that the  
peak must fall within to apply user labels.  
To enter Peak Mass, X Value, or Mass Difference, you  
can type values or right-click-drag over a peak. A Mass  
Tolerance or X Tolerance of 1 is automatically entered  
when you right-click-drag. Charge state is also  
automatically entered when you right-click-drag a  
spectrum peak.  
3
8. Click OK.  
The User Label Setup dialog box reappears.  
9. To change any information, double-click the Label field in  
the User Label Setup dialog box.  
Enter the new values, then click OK.  
Hint: You can sort the label list by any field by clicking  
the column header button (see Figure 3-23 on  
page 3-62).  
10. To save the label settings as a .LBC or .LBS file, click  
Save As, type a file name, then click Save.  
11. Click OK.  
The Peak Label dialog box reappears.  
3-64  
Applied Biosystems  
     
Peak Labeling  
Applying user You can apply labels you previously created and saved in  
.LBC or .LBS files.  
labels from .LBC  
or .LBS files  
Open the .LBC or .LBS file by clicking  
in the User Label  
Setup dialog box. Select the file, then click Open.  
The labels are imported into the .DAT file and displayed in the  
label list. You can modify the settings, delete or add labels,  
and save the changes in a new .LBC or .LBS file.  
NOTE: Changes you make to user labels imported into a  
.DAT file are not saved in the .LBC or .LBS file from which  
they originated. Click Save As, then resave the file if you  
want the changes saved to the originating file.  
Displaying user Select both Enable Labeling and User Labels in the Peak  
Label dialog box to display user labels. Peaks that do not  
meet the peak detection criteria are not labeled.  
labels  
3
NOTE: If you specify user labels with an X Tolerance that is  
outside the acquisition range for the data file, the user  
labels are not displayed. Be aware that the default  
X Tolerance of 1 may be outside the acquisition range when  
the chromatogram is displayed in retention time.  
Data ExplorerSoftware Users Guide  
3-65  
       
Chapter 3 Peak Detection and Labeling  
User labels not If a user label is not displayed or is displayed incorrectly,  
possible causes are:  
displayed  
User labels are not enabled.  
The peak does not fall within the specified Tolerance.  
The label is too long or peaks are too close together. If  
there is not enough room for the label to be displayed, the  
label may be suppressed. Zooming the region of interest  
expands the trace and may allow the labels to be  
displayed.  
To display labels when peaks are close together, select  
Allow overlapping peak labels, or change Orientation  
to Vertical or 45 degrees in the Peak Label dialog box.  
Peak filtering is enabled, and only peaks that meet the  
peak list filtering criteria are labeled.  
Filter Width is set too high to detect the peak.  
If you have Match Charge State enabled for spectrum  
user labels, the peak of interest does not have the charge  
state specified in the user label.  
3
3-66  
Applied Biosystems  
 
Process that Occurs During Peak Detection, Centroiding, and Integration  
3.6 Process that Occurs  
During Peak Detection,  
This section gives an overview of the process that occurs  
during peak detection, centroiding and integration. Parameters  
Peak detection During peak detection, the Data Explorer software:  
Divides the trace into detection ranges based on the  
number of data points per peak (described in  
Section 3.1.2, The Resolution-Based Peak Detection  
Routine).  
Starts scanning the chromatogram or spectrum at the  
3
origin of the x-axis.  
Applies a Gaussian sliding filter to the number of raw data  
points defined by the Filter Width setting. The filter moves  
across the raw data by the number of data points  
specified by the Increment parameter. An Increment of 1  
is used for chromatogram data.  
Searches for an upward-to-downward inflection point  
greater than the larger of the two specified peak detection  
thresholds (% Base Peak Intensity or Minimum Intensity).  
A Minimum Intensity of 2 is used for chromatogram data.  
This inflection point defines the apex region.  
After the apex region is identified, scans to find the  
downward-to-upward inflection points. These downward  
to upward inflection points are the valley regions of the  
peak.  
Data ExplorerSoftware Users Guide  
3-67  
           
Chapter 3 Peak Detection and Labeling  
For spectral data, determines the peak boundaries by  
one of two means:  
If the Noise Threshold is greater than zero, the  
software scans from the valley regions toward the  
apex region using the number of data points  
defined by the Filter Width. If the difference  
between two consecutive filtered regions is  
greater than the Noise Threshold, the midpoint of  
the filter region closest to the apex is used as the  
peak bound.  
If the Noise Threshold is zero, the software scans  
the valley regions for the minimum points to  
determine the peak bounds.  
For chromatogram data, determines the peak boundaries  
by:  
Determining the Noise Threshold by performing an  
automatic signal-to-noise calculation on the tallest  
peak in the chromatogram and using 75 percent of  
the noise value determined as the threshold.  
3
Scanning from the valley regions toward the apex  
region using the number of data points defined by  
the Filter Width. If the difference between two  
consecutive filtered regions is greater than the  
Noise Threshold, the midpoint of the filter region  
closest to the apex is used as the peak bound.  
Compares the area of the new peak to the %Max Peak  
Area and the Minimum Area. A Minimum Area of 2 is  
used for chromatogram data. If the new peak is within the  
specified area settings, the peak is added to the peak list.  
3-68  
Applied Biosystems  
           
Process that Occurs During Peak Detection, Centroiding, and Integration  
After peak After peaks are detected:  
detection  
Centroid mass is calculated for spectral data, then  
modified by Gaussian peak fitting, if it is selected.  
Chromatographic and spectral peaks are integrated.  
Peak lists are generated for chromatograms and spectra.  
Centroiding The Data Explorer software calculates the peak centroid for  
spectral data by:  
Drawing a projected centroid baseline at the  
percentage of the peak height specified by the  
Centroiding% entered. Centroiding is measured  
between the top of the peak and 0. For example, a  
centroiding% of 10 uses the top ten percent of the peak,  
where peak height is determined from 0.  
Searching to the left and right of the apex for data  
points that bound the projected centroid baseline, then  
interpolating the centroid baseline.  
3
Calculating the intensity-weighted area of the peak  
from the left to the right intersection of the centroid  
baseline, then dividing by the area of the same  
region.  
Data ExplorerSoftware Users Guide  
3-69  
       
Chapter 3 Peak Detection and Labeling  
Integration The Data Explorer software integrates chromatographic and  
spectral peaks to calculate the peak area using one of two  
methods to determine the peak baseline (Figure 3-25):  
Valley-to-baselineDrops a vertical line from all valleys  
to a horizontal baseline. The level of the horizontal  
baseline is determined using the minimum peak valley  
point (left or right) for each peak.  
Valley-to-valleyForces a baseline to all valley points.  
3
Valley-to-Baseline  
Valley-to-Valley  
Figure 3-25 Valley-to-Baseline and Valley-to-Valley Integration  
3-70  
Applied Biosystems  
       
Default Peak Detection Settings  
3.7 Default Peak Detection Settings  
This section includes:  
Default .SET files provided  
Additional Voyager .SET files provided  
Default .SET files Default peak detection settings are contained in the following  
.SET files:  
provided  
MARINER.SET  
VOYAGERLINEAR.SET  
VOYAGERREFLECTOR.SET  
VOYAGERPSD.SET  
The following table lists the default settings in .SET files  
provided for chromatograms.  
Chromatogram Settings in All .SET  
Parameter  
3
Files (Mariner and Voyager)  
Basic Settings (chromatogram data)  
Detection Range  
Full  
%Base Peak  
Intensity  
0.25  
%Max Peak Area  
Filter Width  
0.25  
3
Integration  
V-to-B  
Baseline Setting  
Data ExplorerSoftware Users Guide  
3-71  
       
Chapter 3 Peak Detection and Labeling  
The following table lists the default settings in the .SET files  
provided for spectra.  
Voyager Spectrum  
Mariner  
Spectrum  
MARINER.S  
ET  
Reflector  
VOYAGER  
REFLECTOR  
.SET  
Linear  
VOYAGER  
LINEAR.SET  
PSD  
VOYAGER  
PSD.SET  
Parameter  
Basic Settings (spectrum data)  
%Base Peak Intensity  
%Max Peak Area  
0.25  
0
0
0
0.25  
ON  
2
2
2
Use Resolution Dependent  
Settings  
ON  
ON  
OFF  
Mass Resolution  
5,000  
2,000  
10,000  
N/A  
3
Peak Processing  
Integration Baseline Setting V-to-V  
V-to-B  
50  
V-to-B  
50  
V-to-B  
50  
Centroid%  
50  
6
Max. Charge State  
Max. Isotopes#  
Min. Intensity%  
Max. Intensity%  
1
1
1
6
5
10  
5
5
10  
10  
10  
100  
100  
100  
100  
3-72  
Applied Biosystems  
Default Peak Detection Settings  
Voyager Spectrum  
Mariner  
Spectrum  
MARINER.S  
ET  
Reflector  
PSD  
Linear  
VOYAGER  
LINEAR.SET  
Parameter  
VOYAGER  
VOYAGER  
PSD.SET  
REFLECTOR  
.SET  
Advanced  
Detection Range  
Resolution Resolution Resolution Five  
dependent dependent dependent ranges  
Filter Width  
Resolution Resolution Resolution Range  
dependent dependent dependent dependent  
Increment  
Resolution  
dependent  
1
1
1
Noise Threshold  
%Base Peak Intensity  
%Max Peak Area  
Min. Intensity  
0
0
0
2
0
0
0
0
2
0
0
0
0
2
0
3
0.25  
0.25  
0
Min. Area  
0
Range  
dependent  
Additional Additional .SET files that have been developed for detection of  
different types of data are included in the  
C:\VOYAGER\SETTINGS directory. The names of the .SET  
files indicate the type of data the files can be used for.  
Voyager .SET files  
provided  
Data ExplorerSoftware Users Guide  
3-73  
Chapter 3 Peak Detection and Labeling  
3
3-74  
Applied Biosystems  
Chapter  
4
4 Examining  
ChromatogramData  
This chapter contains the following sections:  
4.1  
4.2  
4.3  
Overview ......................................................... 4-2  
Creating an Extracted Ion Chromatogram ........ 4-5  
Creating an Extracted XAC Chromatogram  
(Mariner Data Only) ....................................... 4-13  
4.4  
4.5  
4.6  
Noise Filtering/Smoothing.............................. 4-17  
Adding and Subtracting Spectra .................... 4-20  
Displaying MS Method Data  
(Mariner Data Only) ........................................ 4-23  
4.7  
4.8  
Adjusting the Baseline ................................... 4-27  
Using UV Trace Offset ................................... 4-30  
Data ExplorerSoftware Users Guide  
4-1  
       
Chapter 4 Examining Chromatogram Data  
4.1 Overview  
This section includes:  
Types of Mariner data  
Types of Voyager data  
Creating macros to combine processing functions  
Returning to the original trace  
Types of You can display different types of chromatogram traces from  
Mariner data files by selecting Traces from the Display menu  
with the Chromatogram window displayed, then selecting:  
Mariner data  
Select  
TIC  
To display  
Total Ion Chromatogram, which includes the entire mass range  
saved in the data file. Each data point represents the sum of all ion  
intensities in the corresponding spectrum.  
BPI  
Base Peak Intensity trace, which includes only the base peak in  
each spectrum. Each data point represents the single most intense  
ion in the corresponding spectrum.  
Analog  
Trace of the input from an outside source representing any signal  
that changes over time, for example, the UV signal from an  
LC system, or the Analyzer Temperature, Air Temperature, Spray  
Tip Potential, or Nozzle Temperature from the Mariner mass  
spectrometer.  
4
DAD TAC  
Diode Array Detector (DAD) Total Absorbance Chromatogram  
(TAC), which displays data acquired from LC samples.  
Hint: To display spectra for DAD data, double-click the TAC or  
Ch Absorbance trace.  
DAD  
Channel N  
Diode Array Detector Channel chromatograms, which display  
specific wavelength ranges, where N is the channel number set  
during acquisition. The data file can include up to five channels.  
Hint: To display spectra for DAD data, double-click the TAC or  
Ch Absorbance trace  
4-2  
Applied Biosystems  
                                   
Overview  
You can display extracted chromatograms from Mariner data  
files by selecting the Process menu with a Chromatogram  
window displayed, then selecting:  
Select  
To display  
Extracted Ion—  
Center Window  
or Range  
Extracted Ion Chromatogram (XIC), which includes only the  
signal response from a mass window or range. For more  
Chromatogram (XIC).  
Extracted Ion—  
Neutral Loss  
Constant Neutral Loss chromatogram, which extracts only the  
response from peaks that are separated by a selected mass  
Constant Neutral Loss (CNL) Chromatogram.  
Extracted  
Absorbance  
(XAC)  
Extracted Absorbance Chromatogram, which includes only the  
response from a specified wavelength window or range. For  
more information, see Section 4.3, Creating an Extracted  
Absorbance Chromatogram (XAC) (Mariner Data Only).  
Types of Voyager chromatograms can optionally be displayed for  
multispectrum .DAT files. To display a Voyager chromatogram,  
Voyager data  
select Restore Chromatogram from the View menu. For  
more information, see Section 2.1.3, Displaying Voyager  
Chromatograms.  
4
If you have a chromatogram displayed, you can display the  
following type of Voyager data by selecting Traces from the  
View menu, then selecting:  
Select  
TIC  
To display  
Total Ion Chromatogram, which includes the entire mass range  
saved in the data file. Each data point represents the sum of all ion  
intensities in the corresponding spectrum.  
Data ExplorerSoftware Users Guide  
4-3  
       
Chapter 4 Examining Chromatogram Data  
You can display the following types of Voyager data by  
selecting Extracted Ion from the Process menu with a  
Chromatogram window displayed, then selecting:  
Select  
To display  
Center Window  
or Range  
(XIC)  
from a mass window or range. For more information, see  
Section 4.2.1, Creating an Extracted Ion Chromatogram (XIC).  
Neutral Loss  
(CNL)  
Constant Neutral Loss Chromatogram, which extracts only the  
response from a mass difference from a selected peak. For  
more information, see Section 4.2.2, Creating a Constant  
Neutral Loss (CNL) Chromatogram.  
Creating macros You can create macros that perform multiple functions (for  
example, smooth and baseline-correct), then start the macro  
to combine  
processing  
For information, see Section 6.7, Using the Macro Recorder.  
functions  
Returning to the Many processing functions generate a new trace. If you have  
Trace Replace mode set to Replace, the new trace replaces  
original trace  
the original trace. For information on Replace mode, see  
Section 2.4.4, Adding Traces from the Same Data File to a  
Window.  
To redisplay the original trace, select the trace type from the  
Display menu or click the corresponding toolbar button.  
4
4-4  
Applied Biosystems  
   
Creating an Extracted Ion Chromatogram  
4.2 Creating an  
Extracted Ion Chromatogram  
This section includes:  
Creating an Extracted Ion Chromatogram (XIC)  
Creating a Constant Neutral Loss (CNL) Chromatogram  
You can create an extracted ion chromatogram:  
To improve the signal-to-noise ratio for a mass or mass  
range of interest  
To determine if mass differences in a data file correspond  
to loss of specific fragments by generating a Neutral Loss  
Chromatogram  
4.2.1 Creating an Extracted  
Ion Chromatogram (XIC)  
This section describes how to create an XIC:  
From the Chromatogram window  
From the Spectrum window  
From the To create an extracted ion chromatogram (XIC) from the  
Chromatogram  
4
Chromatogram window:  
window  
1. Click the Chromatogram window to activate it.  
2. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
3. From the Process menu, select Extracted Ion or click  
in the toolbar.  
Data ExplorerSoftware Users Guide  
4-5  
           
Chapter 4 Examining Chromatogram Data  
4. In the Extracted Ion Chromatogram dialog box (Figure 4-1  
on page 4-7), select one of the following from the Mass  
Range/Difference Type drop-down list:  
Center/Window, then type the mass of  
interest and the mass window for masses to  
include.  
NOTE: When analyzing multiple components with  
similar masses, set a Window of less than 0.5 to  
include only the mass of interest.  
Range, then type the From and To values for  
masses to include.  
NOTE: To improve the signal-to-noise ratio in the  
extracted ion chromatogram, set the starting  
NOTE: For more information on Constant Neutral Loss  
chromatograms, see Section 4.2.2, Creating a Constant  
Neutral Loss (CNL) Chromatogram.  
4
4-6  
Applied Biosystems  
   
Creating an Extracted Ion Chromatogram  
Figure 4-1 Extracted Ion Chromatogram Dialog Box  
5. Specify the Extraction Mode:  
AccumulativeCreates one extracted ion  
chromatogram for all masses entered and sums  
intensities  
IndividualCreates one extracted ion  
chromatogram for each mass entered  
6. Click OK.  
The extracted ion chromatogram is displayed in the  
Chromatogram window (Figure 4-2 on page 4-8) with  
an XIC trace label and the center mass and window or  
the mass range indicated in the trace label  
Data ExplorerSoftware Users Guide  
4-7  
     
Chapter 4 Examining Chromatogram Data  
From the To create an extracted ion chromatogram for a mass range  
from the Spectrum window:  
Spectrum window  
1. Click the Chromatogram window to activate it.  
2. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
3. In the Spectrum window, right-click-drag over the mass  
region of interest in the extracted ion chromatogram. The  
width of the box you draw defines the precise mass range  
used in the extracted ion chromatogram.  
The extracted ion chromatogram is displayed in the  
Chromatogram window (Figure 4-2) with the mass  
range indicated in the trace label.  
Extracted ion chromatogram  
(center window)  
4
Figure 4-2 Extracted Ion Chromatogram  
4. To return to the original trace, see Returning to the  
original traceon page 4-4.  
4-8  
Applied Biosystems  
     
Creating an Extracted Ion Chromatogram  
4.2.2 Creating a Constant  
Neutral Loss (CNL) Chromatogram  
This section includes:  
Overview  
Applications  
Labeling spectrum peaks with mass difference (optional)  
Procedure  
Example  
Overview To rapidly screen for the presence of mass differences  
corresponding to loss of specific fragments, you can create a  
Constant Neutral Loss (CNL) extracted chromatogram.  
When you generate a Constant Neutral Loss Extracted ion  
chromatogram, the software constructs a chromatogram that  
contains signal only for spectra that contain peaks separated  
by the specified mass difference. To construct a Neutral Loss  
Extracted ion chromatogram, the software:  
Compares each mass in the peak list to every other mass  
in the peak list and derives mass differences.  
Takes the total signal of any pairs of peaks that generated  
the mass difference and includes it in the chromatogram,  
if any mass difference corresponds to a mass difference  
plus tolerance specified.  
Applications You can generate a CNL extracted chromatogram:  
4
To determine if a fragment of known mass is present  
To increase sensitivity of a trace by including signal from  
both the parent and one or more fragment ions  
For LCMS data, a CNL extracted chromatogram is useful to  
identify eluting components with a given neutral loss.  
For other types of data, a CNL extracted chromatogram is  
useful to screen samples for a given mass difference. For  
example, create a CNL extracted chromatogram to survey a  
peptide digest for the presence of peaks separated by a  
glycosyl residue, as you might see in a digest containing  
glycopeptides.  
Data ExplorerSoftware Users Guide  
4-9  
             
Chapter 4 Examining Chromatogram Data  
Labeling You can label spectrum peaks with mass differences to assist  
you in determining the mass differences to specify in the CNL  
extracted chromatogram.  
spectrum peaks  
with mass  
To label spectrum peaks with mass differences:  
difference  
(optional)  
1. Display a spectrum of interest from the data file for which  
you are creating the CNL extracted chromatogram.  
2. Click the Spectrum window to activate it, then select Peak  
Label from the Peaks menu.  
3. In the Peak Mass Label Type, select Mass difference  
from the selected peak.  
4. In the Spectrum window, right-click-drag across the peak  
from which you are measuring the difference. This enters  
a negative value that corresponds to the reference peak in  
5. Click OK. The reference peak is labeled with zero, and all  
the other peak labels are plus or minus their mass  
difference from the reference peak.  
For more information on peak labels, see Section 3.5,  
Peak Labeling.  
Procedure To create a CNL extracted chromatogram:  
1. Click the Chromatogram window to activate it.  
2. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
4
3. From the Process menu, select Extracted Ion or click  
in the toolbar.  
The Extracted Ion Chromatogram dialog box appears  
(Figure 4-3).  
4. In the Mass Range/Difference Type field select Neutral  
Loss.  
4-10  
Applied Biosystems  
 
Creating an Extracted Ion Chromatogram  
5. Type in the Mass Difference and Tolerance.  
NOTE: Do not right-click-drag across a peak in the  
Spectrum window to select the mass difference.  
Figure 4-3 Extracted Ion Chromatogram Dialog Box  
with Neutral Loss Selected  
4
6. Specify the Extraction Mode:  
AccumulativeCreates one extracted ion  
chromatogram for all mass differences entered and  
sums intensities  
IndividualCreates one extracted ion  
chromatogram for each mass difference entered  
7. Click OK.  
The CNL is displayed in the Chromatogram window  
(Figure 4-5).  
Data ExplorerSoftware Users Guide  
4-11  
   
Chapter 4 Examining Chromatogram Data  
Example Figure 4-4 shows a TIC that contains three flavonoid  
compound peaks. To determine if the diglycosyl group has  
fragmented from the parent ion in any of these compounds,  
you can generate a CNL extracted chromatogram.  
Figure 4-4 TIC for Flavonoid Mixture Containing Three Peaks  
Figure 4-5 shows a CNL extracted chromatogram generated  
for mass difference of 308.146 m/z which corresponds to the  
diglycosyl group.  
4
Figure 4-5 CNL Containing Two Peaks  
The CNL extracted chromatogram contains only two peaks,  
indicating that the diglycosyl group has fragmented from the  
parent ion in the first two compounds only.  
4-12  
Applied Biosystems  
     
Creating an Extracted Absorbance Chromatogram (XAC) (Mariner Data Only)  
4.3 Creating an Extracted  
Absorbance Chromatogram  
(XAC) (Mariner Data Only)  
This section describes how to create an Extracted Absorbance  
chromatogram (XAC) for DAD Mariner data:  
From the Chromatogram window  
From the Spectrum window  
From the To create an extracted absorbance chromatogram (XAC) for a  
selected wavelength window or range:  
Chromatogram  
window  
1. Click the Chromatogram window to activate it.  
2. Display a TAC or Channel chromatogram by selecting  
Traces from the View menu, then selecting the  
chromatogram.  
3. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
4. Select Extracted Absorbance from the Process menu or  
click  
in the toolbar.  
5. In the Extracted DAD Chromatogram dialog box  
(Figure 4-6), click one of the following:  
4
Center/Window, then enter the wavelengths  
of interest (from 190 nm to 950 nm) and the  
window for wavelengths to include.  
Range, then enter the From and To values for  
wavelengths from 190 nm to 950 nm.  
Data ExplorerSoftware Users Guide  
4-13  
         
Chapter 4 Examining Chromatogram Data  
Figure 4-6 Extracted Absorbance Chromatogram  
Dialog Box  
6. Specify the Extraction Mode:  
AccumulativeCreates a single trace  
combining intensities of all specified  
wavelengths  
IndividualCreates one trace for each  
specified wavelength  
4
7. Click OK.  
The extracted absorbance chromatograms are  
displayed in the Chromatogram window with an XAC  
trace label.  
4-14  
Applied Biosystems  
     
Creating an Extracted Absorbance Chromatogram (XAC) (Mariner Data Only)  
From the To create an extracted absorbance chromatogram for a  
wavelength range from the Spectrum window:  
Spectrum window  
1. Click the Chromatogram window to activate it.  
2. From the Display menu, select Traces, then select a  
DAD TAC or DAD Channel chromatogram.  
The TAC or Sig Absorbance trace appears in the  
Chromatogram window.  
3. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
4. Double-click the left mouse button or right-click-drag over  
an area of interest in the Chromatogram window.  
A DAD Spectrum appears in the Spectrum window.  
5. Click the Spectrum window to activate it.  
6. In the Spectrum window, right-click-drag over the region of  
interest. The width of the box you draw defines the precise  
wavelength range used in the extracted absorbance  
chromatogram.  
The extracted absorbance chromatogram is displayed  
in the Chromatogram window (Figure 4-7) with an XAC  
trace label and the center mass and window, or the  
mass range indicated in the trace label.  
4
Data ExplorerSoftware Users Guide  
4-15  
   
Chapter 4 Examining Chromatogram Data  
Figure 4-7 Extracted Absorbance Chromatogram  
7. To return to the original trace, see Returning to the  
original traceon page 4-4.  
4
4-16  
Applied Biosystems  
   
Noise Filtering/Smoothing  
4.4 Noise Filtering/Smoothing  
Description The Noise Filter/Smooth command provides three options for  
reducing noise in chromatogram traces:  
Noise filter  
Smooth by the Gaussian method  
Noise removal  
Procedure To noise filter or smooth a chromatogram trace:  
1. From the Process menu, select Noise Filter/Smooth.  
The Noise Filter/Smooth dialog box (Figure 4-8) is  
displayed.  
Figure 4-8 Noise Filter/Smooth Dialog Box  
4
Data ExplorerSoftware Users Guide  
4-17  
         
Chapter 4 Examining Chromatogram Data  
2. Select the method to use based upon the type of data you  
are examining, then enter the associated value displayed  
for the method you select:  
Suggested  
Method  
Type of Data  
Description  
Higher resolution  
data  
Noise Filter Specify a Correlation Factor of 0 to 1.0.  
(NF)  
Settings from 0.5 to 0.7 yield acceptable  
results for most data. A setting close to 1.0  
(May affect  
peak  
If applying the Noise Filter with a certain  
resolution.) Correlation Factor does not yield the  
necessary noise removal, return to the original  
trace (see Returning to the original traceon  
page 4-4) and apply the Noise Filter again with  
a higher Correlation Factor setting. Applying  
the filter more than one time with the same  
Correlation Factor setting does not improve  
noise removal.  
Higher molecular  
weight data  
Gaussian  
Smooth  
(SM)  
Specify a Filter Width in data points (odd  
integers only). The maximum number of  
smoothing points is 2001. Points less than 1  
Filter Width from the edge of the spectrum are  
not smoothed.  
(May affect  
peak  
4
resolution.)  
4-18  
Applied Biosystems  
   
Noise Filtering/Smoothing  
Suggested  
Method  
Type of Data  
Description  
High-resolution data  
Noise  
Removal  
(NR)  
Specify the number of standard deviations of  
noise to remove. The software automatically  
calculates the average white noise for all  
frequencies across the spectrum, then  
removes the specified number of standard  
deviations of noise. This method slightly  
affects peak intensity, and removes peaks with  
a signal-to-noise ratio less than the specified  
standard deviation.  
(Does not  
affect peak  
resolution.)  
In general:  
If you set  
Std Dev to:  
% White Noise  
removed is:  
1
2
3
68  
99  
If applying Noise Removal with a certain Std  
Dev does not yield the necessary noise  
removal, return to the original trace (see  
Returning to the original traceon page 4-4)  
and apply the Noise Removal again with a  
higher Standard Deviation setting. Applying  
the Noise Removal more than one time with  
the same Standard Deviation setting will not  
improve noise removal.  
4
3. Click OK.  
The trace is displayed with an NF, SM, or NR trace  
label.  
4. To return to the original trace, see Returning to the  
original traceon page 4-4.  
Data ExplorerSoftware Users Guide  
4-19  
     
Chapter 4 Examining Chromatogram Data  
4.5 Adding and Subtracting  
Raw Spectra Within a Data File  
Use the Add/Subtract Spectra command to manipulate raw  
spectra within a single data file.  
NOTE: To manipulate processed spectra, spectra from  
different data files, or spectra acquired under slightly  
different instrument calibrations, use Trace Arithmetic. For  
more information, see Section 5.12, Adding and Subtracting  
Raw or Processed Spectra from the Same or Different Data  
Files (Dual Spectral Trace Arithmetic).  
Adding and To add and subtract spectra:  
subtracting  
spectra  
1. Click the Spectrum window to activate it.  
2. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
3. From the Process menu, select Add/Subtract Spectra.  
The Add and Subtract Spectra dialog box is displayed  
(Figure 4-9).  
4
4-20  
Applied Biosystems  
         
Adding and Subtracting Raw Spectra Within a Data File  
Figure 4-9 .Add and Subtract Spectra Dialog Box  
4. Select spectra to add by doing one of the following:  
Right-click-drag the area of the trace in the  
Chromatogram window. The numbers of the  
selected spectra are added to the list window.  
A range of spectrum numbers is indicated as  
X:X. For example, 10:20 indicates spectrum  
number 10 through spectrum number 20.  
4
Click  
, type the spectrum number range  
to add, for example 10:20, then press Enter.  
Hint: To sum non-contiguous regions of spectra,  
repeat step 4. For example, you can add spectra  
10 to 20 and spectra 30 to 40 to the list window.  
5. Click in the Spectra To Be Subtracted list box.  
Data ExplorerSoftware Users Guide  
4-21  
     
Chapter 4 Examining Chromatogram Data  
NOTE: Before you can subtract spectra, you must first  
specify spectra to be added.  
6. Select spectra to subtract as described in step 4.  
7. Select the Add/Subtract mode:  
AverageSpectra in each list are averaged  
before the addition or subtraction occurs.  
AccumulateSpectra in each list are  
summed before the addition or subtraction  
occurs.  
NOTE: If you select Accumulate mode, include at  
least the same number of spectra in the Subtract  
range that you include in the Add range. For  
example, if you include spectra 775 to 794 in the Add  
range, make sure to include 20 or more spectra in the  
Subtract range.  
8. Click OK.  
The subtracted baseline trace is displayed, with the  
added and subtracted spectrum numbers in the trace  
label. Also, if the retention time labels are displayed,  
the retention time range is included in the trace label.  
4
4-22  
Applied Biosystems  
   
Displaying MS Method Data (Mariner Data Only)  
4.6 Displaying MS Method Data  
(Mariner Data Only)  
Overview If you acquired a data file using an MS Method and assigned  
event tags, you can display chromatogram traces in Data  
Explorer filtered by event tag.  
In this section This section includes:  
Example applications  
Displaying acquisition conditions and event tags  
Setting Trace Replace mode  
Filtering event tags  
Evaluating filtered traces  
Displaying additional filtered traces  
Example Assume that you acquired data using two events, one  
specifying a low Nozzle Potential, and one specifying a higher  
applications  
Nozzle Potential to induce in-source fragmentation. You also  
assigned a Tag 1 event tag to the first event and the In-Source  
CID event tag to the second event.  
You can filter the TIC (or any chromatogram data) to display  
separate traces for the different Nozzle Potentials.  
Displaying To display instrument settings, MS Method segment and event  
numbers, and event tags for a spectrum:  
acquisition  
conditions and  
event tags  
1. Display the spectrum of interest.  
2. Click the Instrument Setting tab in the Output window.  
Setting Trace Before selecting the event tag to display, set the Trace  
Replace mode (Add or Replace) in the Display Trace dialog  
Replace mode  
box. For information on Trace Replace mode, see  
Section 2.4.4, Adding Traces from the Same Data File to a  
Window.  
Data ExplorerSoftware Users Guide  
4-23  
                 
Chapter 4 Examining Chromatogram Data  
Hint: Add mode is useful when you filter the same trace for  
different event tags. The original trace remains displayed  
and accessible. Each filtered trace (up to four total traces)  
is added, allowing for visual comparison.  
Filtering To display chromatogram traces for selected event tags:  
event tags  
1. Display the data file containing the event tags.  
2. Click the Chromatogram window to activate it.  
3. From the Display menu, select Traces, then select the  
trace type to filter.  
4. If you have more than one trace displayed, select the  
trace to filter.  
5. From the Process menu, select Event Tag Filtering.  
NOTE: If the Chromatogram window is not active,  
Event Tag Filtering is not displayed on the Process  
menu. If the active chromatogram was not collected  
with an MS Method that specified event tags, the Event  
Tag Filtering command is dimmed.  
The Event Tag dialog box is displayed (Figure 4-10).  
4
Figure 4-10 Event Tag Dialog Box  
4-24  
Applied Biosystems  
       
Displaying MS Method Data (Mariner Data Only)  
NOTE: Only tags present in the data file are available.  
6. Select the event tags to display, then click OK.  
The filtered trace is displayed with an EF trace label  
(Figure 4-11).  
Figure 4-11 Event Filtered Trace  
NOTE: The spectra in an event-filtered trace are  
numbered contiguously (1,2,3...) regardless of their  
relation to the overall acquisition. However, because  
the axes of the trace reflect the numbering of the  
overall experiment, you may see spectra with numbers  
that do not correspond to the x-axis.  
4
Hint: Line mode is useful when displaying filtered LC  
TIC traces. Vertical bar mode may be more useful when  
filtering direct infusion TIC traces.  
If you have the Replace mode set to Add in the Display  
Trace dialog box, a new trace is added.  
Data ExplorerSoftware Users Guide  
4-25  
         
Chapter 4 Examining Chromatogram Data  
Hint: Add mode is useful when you are filtering the  
remains displayed and accessible. Each filtered trace  
(up to four total traces) is added, allowing for visual  
comparison.  
7. To return to the original trace, see Returning to the  
original traceon page 4-4.  
Evaluating Note the following when evaluating filtered traces:  
filtered traces  
If you select more than one tag to display, all spectra  
containing the specific combination of tags are included in  
the filtered trace.  
For example, assume the MS Method used to acquire the  
data specifies Tag 1 for event 1, and both Tag 1 and Tag 2  
for event 2. If you select Tag 1 to filter, spectra from  
event 1 are included in the filtered trace, but not spectra  
from event 2.  
A filtered trace displays data points for the selected tags  
only. However, if you double-click the filtered trace in an  
area that does not appear to contain data, a spectrum is  
displayed in the Spectrum window.  
Any actions you perform on a filtered trace, such as  
summing, include only the spectra displayed in the  
filtered trace. Spectra from areas of the filtered trace that  
do not appear to contain data are not included.  
4
NOTE: In a summed trace, the trace label includes the  
spectrum numbers for all spectra in the summing  
range, even if they are not in the filtered trace.  
However, only the spectra with the appropriate tags are  
included in the summed spectrum.  
Displaying To display additional filtered traces:  
additional filtered  
1. Select the original, unfiltered trace.  
traces  
2. From the Process menu, select Event Tag Filtering, then  
select the tags to display.  
4-26  
Applied Biosystems  
 
Adjusting the Baseline  
4.7 Adjusting the Baseline  
This section includes:  
Using Baseline Offset  
Using Baseline Correction  
4.7.1 Using Baseline Offset  
Use the Baseline Offset command to offset the y-axis in a  
chromatogram to improve the appearance of a trace or correct  
a sloping baseline.  
To use Baseline Offset:  
1. Activate the window on which you want to perform the  
offset.  
NOTE: You can select a Chromatogram or Spectrum  
window. If you do not activate the correct type of  
window before performing the next step, the software  
does not select values when you click-drag on the  
trace. For example, you must select a Chromatogram  
window before starting baseline offset on a  
chromatogram.  
4
2. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
3. From the Process menu, select Baseline Offset.  
The Baseline Offset dialog box (Figure 4-12) is  
displayed.  
Data ExplorerSoftware Users Guide  
4-27  
       
Chapter 4 Examining Chromatogram Data  
Figure 4-12 Baseline Offset Dialog Box  
4. Right-click-drag the left baseline to offset. The selected  
value is displayed in the Left Baseline field.  
5. Right-click-drag the right baseline to offset. The selected  
value is displayed in the Right Baseline field.  
6. To limit the baseline offset to the area between the two  
selected points, select Only Apply from L to R Baseline  
Midpoint.  
To perform the baseline offset on the entire x-axis,  
deselect Only Apply from L to R Baseline Midpoint.  
7. Click OK.  
The offset baseline trace is displayed with a BO trace  
label.  
4
8. To return to the original trace, see Returning to the  
original traceon page 4-4.  
4-28  
Applied Biosystems  
   
Adjusting the Baseline  
4.7.2 Using Baseline Correction  
Description The Baseline Correction feature is a function that corrects for  
a curved baseline, including a DC-offset baseline, by  
eliminating broad artifacts from the data set.  
When to use Baseline-correct if you are analyzing data:  
With a baseline that is not flat, and you are using the  
%Base Peak Intensity parameter (intensity-based  
thresholding) to screen out noise peaks. For best results,  
apply Baseline Correction, then re-detect peaks.  
If you use area-based thresholding (%Max Peak Area)  
during peak detection, Baseline Correction is not typically  
needed. %Max Peak Area compensates for a rising or  
falling baseline.  
With a baseline that is not at 0, to improve peak detection  
of small peaks.  
Correcting the To correct the baseline:  
baseline  
1. Display the spectrum of interest.  
2. From the Process menu, select Baseline Correction.  
The baseline is adjusted, and the trace is displayed  
with a BC trace label.  
Returning to the To return to the original trace, see Returning to the original  
4
traceon page 4-4.  
original trace  
Data ExplorerSoftware Users Guide  
4-29  
         
Chapter 4 Examining Chromatogram Data  
4.8 Using UV Trace Offset  
(Mariner Data Only)  
To align a UV trace with a chromatogram trace:  
1. Display the chromatogram and UV traces of interest.  
2. Display the x-axis in retention time for each trace by  
selecting Traces from the Display menu, selecting X Axis  
In, then selecting Time.  
3. Set the Replace Mode to Add a New Trace. For  
information see Setting the Replace modeon page 2-17.  
4. From the Process menu, select Realign UV Trace.  
The UV Trace Offset dialog box is displayed  
(Figure 4-13).  
Figure 4-13 UV Trace Offset Dialog Box  
5. Enter the retention times as follows.  
4
Click the Original UV Peak Value (min) text box, then  
right-click-drag over the peak to offset in the UV trace.  
Select the Aligned UV Peak To (min) text box, then  
right-click-drag over the peak to align with in the  
chromatogram trace.  
Hint: You can type retention times in the UV Trace  
Offset dialog box. Use an X data cursor to read the  
retention time of the peak. If you enter spectrum  
numbers by mistake, the peak does not align properly.  
4-30  
Applied Biosystems  
       
Using UV Trace Offset (Mariner Data Only)  
6. Click OK.  
The UV trace peak is shifted to align with the  
chromatogram trace peak.  
NOTE: To restore the original UV trace, open the UV  
Trace Offset dialog box (see step 4), then click Reset.  
7. To return to the original trace, see Returning to the  
original traceon page 4-4.  
4
Data ExplorerSoftware Users Guide  
4-31  
 
Chapter 4 Examining Chromatogram Data  
4
4-32  
Applied Biosystems  
Chapter  
5
5 Examining  
Spectrum Data  
This chapter contains the following sections:  
5.1  
5.2  
5.3  
5.4  
5.5  
5.6  
5.7  
5.8  
5.9  
Overview ......................................................... 5-2  
Creating a Combined Spectrum ...................... 5-4  
Manual Calibration........................................... 5-5  
Automatic Calibration..................................... 5-26  
Centroiding .................................................... 5-36  
Mass Deconvolution (Mariner Data Only)....... 5-37  
Noise Filtering/Smoothing.............................. 5-42  
Adjusting the Baseline ................................... 5-45  
Truncating a Spectrum................................... 5-56  
5.10 Converting to a Singly Charged Spectrum  
(Mariner Data Only) ....................................... 5-59  
5.11 AutoSaturation Correction  
(Mariner Data Only) ....................................... 5-62  
5.12 Adding and Subtracting Raw or  
Processed Spectra From the Same or  
Different Data Files  
(Dual Spectral Trace Arithmetic) .................... 5-64  
Data ExplorerSoftware Users Guide  
5-1  
       
Chapter 5 Examining Spectrum Data  
5.1 Overview  
Types of spectra You can display the following types of spectrum data:  
you can display  
Single spectrumDouble-click any point in the TIC to  
display the corresponding spectrum.  
Combined spectrumClick-drag across a region of  
the chromatogram to sum the intensity at each mass for  
all selected spectra. Combining spectra can improve  
peak shape, signal intensity, and signal-to-noise ratio.  
NOTE: You cannot display a combined spectrum for  
Voyager single-spectrum data files.  
DAD spectrum (Mariner DAD data only)Click the  
Chromatogram window, select Traces from the Display  
menu, select the DAD TAC or channel of interest, then  
double-click any point in the TAC or channel data trace to  
display the corresponding DAD spectrum.  
Voyager PSD Voyager PSD data files contain the precursor spectrum and  
fragment ion spectra. You can view up to eight spectra at a  
spectra  
time and can click  
fragment ions.  
and  
to display other  
Creating macros You can create macros that perform multiple functions (for  
example, smooth and baseline correct), and then start the  
macro with one mouse click.  
to combine  
processing  
For information, see Section 6.7, Using the Macro Recorder.  
functions  
5
5-2  
Applied Biosystems  
                   
Returning to the Many processing functions generate a new trace. If you have  
Trace Replace mode set to Replace, the new trace replaces  
original spectrum  
the original trace. For information on Replace mode, see  
Section 2.4.4, Adding Traces from the Same Data File to a  
Window.  
To return to the original spectrum:  
If the original spectrum was an unprocessed spectrum,  
select Spectrum # from the Display menu. The number  
of the original spectrum is displayed in the Select  
Spectrum dialog box. Click OK.  
If the original trace was a processed spectrum, select  
Processing History from the Display menu, then select  
the original trace.  
5
Data ExplorerSoftware Users Guide  
5-3  
   
Chapter 5 Examining Spectrum Data  
5.2 Creating a Combined Spectrum  
NOTE: Before creating a combined spectrum for Voyager  
multispectrum data files, calibrate the data. See  
Section 5.3, Manual Calibration.  
To create a combined spectrum:  
1. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
2. In the Chromatogram window, right-click-drag across the  
region of the chromatogram that contains the spectra to  
combine.  
The combined spectrum (Figure 5-1) is displayed.  
Combined spectrum label  
Figure 5-1 Combined Spectrum  
3. To return to the original trace, see Returning to the  
original spectrumon page 5-3.  
5
5-4  
Applied Biosystems  
           
Manual Calibration  
5.3 Manual Calibration  
This section describes:  
Overview of manual calibration  
Manually calibrating  
Creating or modifying a calibration reference file  
Reverting to instrument calibration  
Hints for calibrating Mariner data  
Hints for calibrating Voyager data  
NOTE: Manual calibration is not supported for Mariner  
DAD data.  
5.3.1 Overview of Manual Calibration  
Overview During manual calibration:  
You specify a calibration reference file (.REF) that  
contains reference masses.  
The software matches peaks by comparing observed  
masses in the spectrum to reference masses.  
The software lists masses that match within the specified  
peak matching criteria, and calculates calibration  
constants.  
You can apply the calibration constants to all spectra in a  
Mariner or Voyager data file.  
You can apply individual calibration constants to each  
spectrum in a Voyager data file.  
5
Data ExplorerSoftware Users Guide  
5-5  
                 
Chapter 5 Examining Spectrum Data  
The manual calibration feature provides two modes for peak  
matching:  
AutomaticThe software automatically compares  
reference masses to observed masses, and lists peaks  
that are within the specified peak matching criteria.  
ManualYou manually select an observed mass, then  
select the reference mass for comparison.  
Hint: It is often useful to automatically perform a match and  
fit first, then manually adjust the fit as needed.  
Calibration Default calibration reference files are provided with the  
software. You can use these files as a starting point and add  
delete reference masses as needed.  
references files  
(.REF) provided  
For information, see Section 5.3.3, Creating or Modifying a  
Calibration Reference File (.REF).  
When to use Use manual calibration when you:  
manual  
Calibrate Voyager data  
calibration  
Calibrate Mariner data and any of the following occur:  
You have one or only a few spectra to calibrate  
You do not know in advance which reference  
masses to use  
You do not know in advance if the quality of the  
You want to fine-tune the results of an automatic  
calibration  
For information on automatic calibration, see Section 5.4,  
Automatic Calibration.  
Accurate mass Before performing an internal calibration, refer to the:  
measurements  
Mariner Workstation Users Guide  
5
Voyager Biospectrometry Workstation Users Guide  
5-6  
Applied Biosystems  
           
Manual Calibration  
5.3.2 Manually Calibrating  
This section describes:  
Before calibrating Voyager data  
Manually calibrating a single spectrum  
Applying new constants to the data file  
Exporting calibration constants (.CAL file)  
Applying new constants to additional files  
Before calibrating Before you calibrate Voyager data, do the following to improve  
mass accuracy:  
Voyager data  
Baseline correctThe Centroid% peak detection value  
is derived from a percentage of the peak height, which is  
measured from 0 (not from the local baseline). For  
information, see Section 5.8.2, Using Baseline  
Correction.  
Noise filter or use default smoothingUse the method  
appropriate for your data to remove noise spikes. For  
more information, see Section 5.7, Noise  
Filtering/Smoothing.  
It is critical to perform both these functions, which affect the  
peak centroid, before calibration. It is also good practice to  
perform the same processing functions on calibrants and  
unknowns.  
NOTE: If you are calibrating Mariner data, baseline  
correction and noise filtering or smoothing are not  
recommended before calibrating. Due to the shorter flight  
times and fewer data points associated with Mariner data,  
these functions may affect peak shape, which in turn affect  
mass accuracy.  
5
Data ExplorerSoftware Users Guide  
5-7  
       
Chapter 5 Examining Spectrum Data  
Manually  
calibrating a  
single spectrum  
NOTE: Multi-point calibration yields higher mass accuracy  
than one-point calibration. Selecting calibrant peaks that  
bracket the mass of interest also yields higher mass  
accuracy.  
To manually calibrate a single spectrum:  
1. Click the Spectrum window to activate it, then select  
the spectrum trace of interest.  
2. From the Peaks menu, select Peak Label, set the Mass  
Label Type to Centroid, then click OK.  
NOTE: For spectra containing broad peaks that have  
unresolved adducts or impurities such as proteins, you  
may obtain better results if you use apex instead of  
centroid settings.  
3. From the Process menu, select Mass Calibration, then  
select Manual Calibration.  
The Manual Mass Calibration dialog box is displayed  
(Figure 5-2).  
5
5-8  
Applied Biosystems  
       
Manual Calibration  
Figure 5-2 Manual Mass Calibration Dialog Box  
4. Select a calibration reference file. For information on  
creating a reference file, see Section 5.3.3, Creating or  
Modifying a Calibration Reference File (.REF).  
5. Enter reference matching and calibration criteria:  
Minimum IntensityPeaks must be above this  
intensity to be considered a match. Select the unit  
for Minimum Intensity, % Relative Intensity or  
Absolute Counts.  
Mass TolerancePeaks must be within this  
tolerance of the theoretical m/z to be considered a  
match. Select the unit, m/z or ppm.  
5
Data ExplorerSoftware Users Guide  
5-9  
     
Chapter 5 Examining Spectrum Data  
6. Select the Peak Weighting Factor. If the calibration  
includes more than two points, you can apply the following  
weighting factors to fit points to the curve:  
NoneAll peaks weighted equally  
Inverse WidthNarrower peaks are weighted  
more than broader peaks  
HeightMore intense peaks are weighted more  
than less intense peaks  
Manually matching 7. To manually select the reference mass for a peak,  
right-click-drag over the peak of interest.  
peaks  
The Select or Create Reference Peak Information  
dialog box (Figure 5-3) is displayed and lists all masses  
in the selected reference file. The selected entry is the  
nearest match in the calibration reference file to the  
selected peak.  
5
Figure 5-3 Select or Create Reference Peak  
Information Dialog Box  
5-10  
Applied Biosystems  
         
Manual Calibration  
8. Do any of the following:  
Click OK to accept the selected reference mass  
for matching, then add it to the Peaks Matched  
list.  
Select a different reference mass, then click OK.  
Type new reference mass information in the  
Name, Theoretical m/z, Charge, and Elemental  
Composition text boxes, select the mass type,  
then click OK to accept the reference mass for  
matching.  
NOTE: You must type a minus sign () preceding  
the charge in the Charge text box for negative  
charge states.  
Type new reference mass information in the  
appropriate text boxes, click Save or Save As to  
add the information to the reference file, then click  
OK to accept the reference mass for matching.  
Type new reference mass information in the  
appropriate text boxes, click Update to replace an  
to add the information to the reference file, then  
click OK to accept the reference mass for  
matching.  
The Manual Mass Calibration dialog box is displayed  
again (see Figure 5-2 on page 5-9) with the observed  
mass and the reference mass you selected displayed in  
the Peaks Matched list.  
Hint: You can sort the list of matches by clicking the  
column header buttons. You can display complete  
information about a reference mass by double-clicking  
the mass.  
5
Data ExplorerSoftware Users Guide  
5-11  
   
Chapter 5 Examining Spectrum Data  
9. Repeat step 7 and step 8 until all desired peaks are in the  
matched list.  
Eliminating data 10. To eliminate unacceptable data points from the  
calibration, do either of the following:  
points  
Select the data point (mass) in the Peaks matched  
list, then click Delete Selected Match.  
Click Eliminate Fit Outlier. The software deletes  
from the calibration the data point with the largest  
calibrated Fit Error (difference between the  
calibrated mass and the reference mass) as  
reported in the Output window (Figure 5-4).  
NOTE: The point deleted may not be the point  
with the largest Initial Error (difference between the  
pre-calibration observed mass and the reference  
mass) listed in the Peaks Matched List in the  
Manual Calibration dialog box.  
5
5-12  
Applied Biosystems  
     
Manual Calibration  
Eliminate Fit Outlier removes the mass associated with the largest fit error  
in the Output window, not the mass associated with the largest initial error  
in the Peaks Matched list.  
The 267 Da mass is  
removed when you  
click Eliminate Fit  
Outliers because  
it generated the  
largest fit error (below).  
Figure 5-4 Eliminate Fit Outlier Deletes the Match with the Largest Fit Error from  
the Output Window  
To clear the entire list, click Delete Entire List.  
Plotting 11. To apply the calibration constants to the displayed  
spectrum, click Plot.  
The spectrum is calibrated and displayed with an  
MC trace label. The calibration statistics are displayed  
in the Result tab of the Output window (Figure 5-5).  
5
Data ExplorerSoftware Users Guide  
5-13  
           
Chapter 5 Examining Spectrum Data  
Calibration  
statistics  
Figure 5-5 Calibration Statistics in Output Window  
If you calibrate more than once, subsequent calibration  
statistics are added to the end of the list in the Output  
window. Older calibration statistics are listed at the top  
of the list.  
Automatically You can automate peak matching by clicking Match Peaks  
instead of right-click-dragging individual peaks and selecting  
the mass.  
matching peaks  
If any masses within the tolerances of any of the masses listed  
in the calibration reference file are found in the spectrum, the  
matches are displayed in the Calibration Mass Peak Selection  
window with an Initial Error reported.  
5
5-14  
Applied Biosystems  
     
Manual Calibration  
NOTE: If you are calibrating Mariner data, see Ensuring  
that masses match during calibrationon page 5-24.  
Applying new To apply the calibration constants to the data:  
constants to the  
data file  
If you are  
calibrating  
Click  
The following occurs  
Mariner data Apply  
All spectra in the data file are calibrated and  
displayed with an MC trace label. The calibration  
constants are saved with the data file. Each  
spectrum in the data file is calibrated when  
displayed.  
(including  
MS Method  
data)  
Calibration  
Voyager data Apply  
The current spectrum is calibrated and  
displayed with an MC trace label. The calibration  
constants are saved with the spectrum.  
Calibration  
Apply to All  
All spectra in the data file are calibrated using  
the currently displayed calibration, and are  
displayed with an MC trace label. The calibration  
constants are saved with the data file. Each  
spectrum in the data file is calibrated when  
displayed.  
NOTE: This  
button is  
displayed only if  
you are  
calibrating a  
Voyager  
multispectrum  
data file.  
NOTE: If you select Apply to All, you overwrite  
any calibrations previously applied to individual  
spectra by clicking Apply Calibration.  
5
Data ExplorerSoftware Users Guide  
5-15  
 
Chapter 5 Examining Spectrum Data  
Exporting The calculated calibration constants can be exported to a  
.CAL file for use with other data files. You have two options for  
exporting:  
calibration  
constants  
(.CAL file)  
Export the current calibrationClick Export in the  
Manual Mass Calibration dialog box to calculate and  
export the current calibration constants from the masses  
displayed in the Peaks Matched list in the Manual Mass  
Calibration dialog box.  
NOTE: The calibration constants saved in the .CAL file  
are not calculated from the active trace on the screen or  
taken from the calibration in the data file. They are  
calculated from the listed masses.  
Export a calibration from a previously calibrated data  
fileOpen an existing data file, then export the  
calibration constants in a .CAL file. See Exporting .BIC,  
.MSM, and .CAL fileson page 1-36.  
Applying new You can apply calibration constants from a .CAL file to any  
data file. To apply the new constants from a mass calibration  
file to a different file:  
constants to  
additional files  
1. Display the spectrum to calibrate.  
2. From the Process menu, select Mass Calibration.  
3. Select Import Calibration.  
NOTE: If you are importing a calibration into a Voyager  
data file, see Importing a calibrationon page 5-25.  
4. Select the .CAL file to use, then click Open.  
The software displays the calibrated spectrum with an  
MC trace label.  
5
5-16  
Applied Biosystems  
           
Manual Calibration  
5. To save the calibration to the data file, select Mass  
Calibration from the Process menu, then:  
If you are  
calibrating  
Select  
The following occurs  
Mariner  
data  
(including  
MS Method  
data)  
Apply Calibration  
All spectra in the data file are calibrated and  
displayed with an MC trace label. The  
calibration constants are saved with the data  
file. Each spectrum in the data file is  
calibrated when displayed.  
Voyager  
data  
Apply Calibration  
Apply to All  
The current spectrum is calibrated and  
displayed with an MC trace label. The  
calibration constants are saved with the  
spectrum.  
All spectra in the data file are calibrated  
using the currently displayed calibration, and  
are displayed with an MC trace label. The  
calibration constants are saved with the data  
file. Each spectrum in the data file is  
calibrated when displayed.  
NOTE: This button is  
displayed only if you  
are calibrating a  
Voyager  
multispectrum data  
file.  
5.3.3 Creating or Modifying a  
Calibration Reference File (.REF)  
This section includes:  
Definition  
.REF files provided  
Calibration reference file contents  
Creating and saving a calibration reference file  
Modifying a calibration reference file  
Specifying mass type for highly charged, narrow peaks  
5
Data ExplorerSoftware Users Guide  
5-17  
         
Chapter 5 Examining Spectrum Data  
Definition A calibration reference file (.REF) is a list of masses and  
corresponding information from which you can select  
reference masses during calibration.  
.REF files The following default calibration reference files are provided  
with the software:  
provided  
MARINER_POS.REF  
MARINER_NEG.REF  
VOYAGER.REF  
ANGIOTENSIN_FRAGMENTS.REF  
IMMONIUM_IONS.REF  
Calibration A calibration reference file is a tab-delimited text file that  
contains the following columns of information:  
reference  
file contents  
Reference mass  
Charge (defaults to 0 if you do not specify)  
TypeResolved, Average, or Unknown (defaults to  
Resolved if you do not specify)  
Name (optional)  
Elemental composition (optional)  
NOTE: If you view the reference file in a text editor (such as  
Notepad Editor), the order of the information listed in the  
reference file is different than the order of the information  
on page 5-10).  
Creating and NOTE: You can also create a reference file using the Ion  
Fragmentation calculator. For more information, see  
Creating a calibration reference file (.REF)on page 6-30.  
saving a  
calibration  
reference file  
To create and save a calibration reference file:  
5
1. From the Process menu, select Mass Calibration, then  
select Edit/Create Reference File.  
5-18  
Applied Biosystems  
                       
Manual Calibration  
The Edit/Create Reference Peak Information dialog box  
(Figure 5-6) is displayed.  
Figure 5-6 Edit/Create Reference Peak Information  
Dialog Box  
2. Type the Name and Theoretical m/z for a reference  
compound, then select the charge state. Optionally, enter  
3. Specify the mass type (Resolved Isotope Mass or  
Average Mass).  
If you are adding masses for highly charged  
compounds, see Specifying mass type for highly  
charged, narrow peakson page 5-21.  
4. Click Insert.  
5
5. Repeat step 2 through step 4 for remaining compounds.  
Data ExplorerSoftware Users Guide  
5-19  
     
Chapter 5 Examining Spectrum Data  
CAUTION  
The software allows you to add multiple items with the  
same m/z value to the reference file if any other  
attribute of the reference compound is different (for  
example, charge state or name). Each mass in the list  
is considered during calibration. If the mass list  
contains duplicate entries, the calibration may return an  
invalid number of matches.  
6. To modify an entry, click the reference mass in the list to  
select it, modify the entry as needed, then click Update.  
The modified entry is saved in the .REF file.  
7. Click Save.  
8. In the Save As dialog box, select a location and type a  
name for the file, then click OK.  
The name of the reference file used for mass calibration  
is stored in the .DAT file and is displayed in the  
Calibration Reference File field when you open the  
Manual Calibration dialog box again.  
NOTE: Only the name and path of the reference file are  
stored in the .DAT file. Contents of the file are not  
stored.  
Modifying a Hint: You can also modify a calibration reference file when  
you manually select reference masses during calibration.  
See step 8 on page 5-11.  
calibration  
reference file  
To add or delete information in a calibration reference file:  
1. From the Process menu, select Mass Calibration, then  
select Edit/Create Reference File.  
The Edit/Create Reference Peak Information dialog box  
(Figure 5-6 on page 5-19) is displayed.  
5
2. Click Browse, then select a calibration reference file.  
5-20  
Applied Biosystems  
     
Manual Calibration  
3. To modify an entry, click the entry to select it, modify the  
entry as needed, then click Update.  
4. To delete an entry, click the entry to select it, then click  
Delete.  
5. To add an entry, type the Name and Theoretical m/z for a  
reference compound, then select the charge state.  
Optionally, enter the Elemental Composition for the  
compound.  
6. Specify the mass type (Resolved Isotope or Average).  
7. Click Insert.  
CAUTION  
The software allows you to add multiple items with the  
same m/z value to the calibration list box if any other  
attribute of the reference compound is different (for  
example, charge state or name). Each mass in the list  
is considered during calibration. If the mass list  
contains duplicate entries, the calibration may return an  
invalid number of matches.  
8. Click Save or Save As.  
9. In the Save As dialog box, select a location and type a  
name for the file, then click Save.  
Specifying mass When specifying highly charged, non-isotopically resolved  
species with peaks less than 1 Da wide (for example,  
myoglobin +20) as reference masses in a calibration reference  
file, set the peak type as a Resolved Isotope Mass (even  
though it is not a resolved isotope). The calibration routine  
type for highly  
charged, narrow  
peaks  
checks peak width to determine if a peak matches a Resolved  
Isotope Mass or an Average mass. If narrow peaks are  
specified as Average Masses in the calibration reference file,  
the software mistakes these narrow peaks as isotopically  
resolved and ignores the reference mass.  
5
Data ExplorerSoftware Users Guide  
5-21  
     
Chapter 5 Examining Spectrum Data  
5.3.4 Reverting to Instrument Calibration  
The Revert to Instrument Calibration function does the  
following:  
Mariner dataReapplies the original calibration  
constants used to acquire the data.  
Voyager dataApplies default calibration to the data,  
regardless of whether default or external calibration is  
used to acquire the data.  
To revert the calibration:  
1. From the Process menu, select Mass Calibration, then  
select Revert to Instrument Calibration.  
The spectrum is recalibrated with the original  
calibration constants used during acquisition (Mariner  
data) or default calibration (Voyager data) and  
displayed with an MC trace label. The original and new  
constants are displayed in the Output window.  
2. To save the reverted calibration with the data file, select  
Mass Calibration from the Process menu, then:  
If you are  
calibrating  
Select  
The following occurs  
Mariner  
data  
(including  
MS Method  
data)  
Apply Calibration  
All spectra in the data file are calibrated and  
displayed with an MC trace label. The  
calibration constants are saved with the data  
file. Each spectrum in the data file is  
calibrated when displayed.  
5
5-22  
Applied Biosystems  
             
Manual Calibration  
If you are  
calibrating  
Select  
The following occurs  
Voyager  
data  
Apply Calibration  
The current spectrum is calibrated and  
displayed with an MC trace label. The  
calibration constants are saved with the  
spectrum.  
Apply to All  
All spectra in the data file are calibrated  
using the currently displayed calibration, and  
are displayed with an MC trace label. The  
calibration constants are saved with the data  
file. Each spectrum in the data file is  
calibrated when displayed.  
NOTE: This button is  
displayed only if you  
are calibrating a  
Voyager  
multispectrum data  
file.  
If you apply the calibration, the next time you open the  
data file, the MC trace label is not displayed.  
5
Data ExplorerSoftware Users Guide  
5-23  
 
Chapter 5 Examining Spectrum Data  
5.3.5 Hints for Calibrating Mariner Data  
Ensuring that Mariner TOF Analyzer parameters affect flight times of ions. If  
you acquired Mariner data using different TOF Analyzer  
parameters, and did not calibrate the data in the Instrument  
masses match  
during calibration  
Control Panel to compensate for the altered parameters,  
masses in the data file may be significantly different from the  
reference masses. For information on calibrating in the  
Instrument Control Panel, see the Mariner Workstation  
Users Guide.  
Before performing automatic matching, set the Mass  
Tolerance and Minimum Intensity appropriately to match the  
expected reference masses. After automatic matching, check  
the list of matched peaks to make sure the correct reference  
masses and peak masses are selected.  
5
5-24  
Applied Biosystems  
   
Manual Calibration  
5.3.6 Hints for Calibrating Voyager Data  
Importing a If you import a calibration, you must import a calibration  
generated from a data file that was acquired on the same  
instrument using identical settings for the following instrument  
setting parameters:  
calibration  
Polarity  
Instrument mode  
If you import a calibration that was generated using settings  
that are different from the current Polarity and Instrument  
Mode settings, an error message is displayed.  
NOTE: The calibration of the mass scale changes when  
you change the Accelerating Voltage, Grid Voltage%, or  
Delay Time. Default calibration adjusts for these changes.  
However, you observe more accurate calibration if you use  
an external calibration (.CAL) file generated with the same  
Accelerating Voltage, Grid Voltage%, or Delay Time.  
5
Data ExplorerSoftware Users Guide  
5-25  
       
Chapter 5 Examining Spectrum Data  
5.4 Automatic Calibration  
This section includes:  
Overview of automatic calibration  
Importing and specifying automatic calibration settings  
Automatically calibrating (Mariner data only)  
NOTE: Automatic calibration is not supported for  
Mariner DAD data.  
5.4.1 Overview of Automatic Calibration  
This section includes:  
During automatic calibration  
Automatic calibration for Mariner data  
Automatic calibration for Voyager data  
During automatic During automatic calibration:  
calibration  
You specify auto calibration settings (reference masses,  
matching criteria, and fit rejection parameters). Auto  
calibration settings are saved as part of processing  
settings in a .DAT file.  
If Auto Calibrate is enabled in the Data Explorer software  
(Mariner data only), or if the Automatic Calibration  
function is accessed by the Voyager Sequence Control  
Panel (Voyager data only), the software does the  
following for each spectrum it displays or processes:  
Compares peaks in the trace to peaks listed in the  
Masses to Match list in auto calibration settings.  
5
5-26  
Applied Biosystems  
         
Automatic Calibration  
Matches all peaks that meet the specified  
Reference Matching criteria. If the number of  
peaks matched is greater than or equal to the  
specified Minimum Number of Peaks to Match,  
and the resulting fit errors are less than or equal to  
the specified Max Outlier Error, calibration is  
successful.  
If any points exceed the specified Max Outlier  
Error, the software eliminates the outliers,  
one-by-one, worst to best, until all points are  
within the specified Max Outlier Error.  
If the number of matching peaks falls below the  
Minimum Number of Peaks to Match when the  
software eliminates an outlier, the calibration fails.  
Automatic The Automatic Calibration function in Data Explorer is useful  
for quickly calibrating all spectra in a Mariner data file, after  
calibration for  
Mariner data  
you prepare automatic calibration settings. You use the  
Automatic Calibration function in Data Explorer to prepare  
Automatic Calibration settings (contain reference masses and  
other matching information) and to calibrate the data.  
You can also use the Automatic Calibration function in Data  
Explorer to prepare Automatic Calibration settings (contain  
reference masses and other matching information) for use by  
the Mariner Sequence Control Panel (described in the Mariner  
Workstation Version 4.0 Supplement). You can save a .SET  
file that contains the settings, and specify the .SET file in the  
Sequence Control Panel. If you create and specify a macro to  
automatically calibrate, the Sequence Control Panel can use  
the calibration settings contained in the .SET file.  
5
Data ExplorerSoftware Users Guide  
5-27  
       
Chapter 5 Examining Spectrum Data  
When to use Use automatic calibration for Mariner data when you:  
Have many spectra to calibrate  
Know in advance what reference masses to use  
Know in advance that the quality of the reference mass  
signals is acceptable  
For information on manual calibration, see Section 5.3,  
Manual Calibration.  
Automatic The Automatic Calibration function in Data Explorer is used by  
the Voyager Sequence Control Panel (described in the  
Voyager Biospectrometry Workstation Users Guide) to  
calibration for  
Voyager data  
calibrate data as it is acquired. You use the Automatic  
Calibration function in Data Explorer to prepare Automatic  
Calibration settings (contain reference masses and other  
matching information). You then save a .SET file that contains  
the settings, then specify the .SET file in the Sequence Control  
Panel. The Sequence Control Panel automatically calibrates  
spectra using the calibration settings contained in the .SET  
file.  
When to use Use automatic calibration for Voyager data to prepare  
Automatic Calibration settings. The Sequence Control Panel  
uses Automatic Calibration settings to calibrate data as it is  
acquired. For more information on the Sequence Control  
Panel and on specifying automatic calibration, see the  
Voyager Biospectrometry Workstation Users Guide.  
5
5-28  
Applied Biosystems  
         
Automatic Calibration  
5.4.2 Importing and Specifying  
Automatic Calibration Settings  
Importing  
Hint: Importing automatic calibration settings is useful  
when you calibrate batches of related samples.  
Automatic calibration settings are saved as part of processing  
settings in a .DAT file. To use auto calibration settings from  
another .DAT file:  
2. Save the .SET information from this file by selecting  
Settings from the File menu, then selecting Save  
Processing Settings As. For more information, see  
Section 1.6.5, Extracting and Saving Information from  
3. Open or activate the new file.  
4. Copy the settings to the new file by selecting Settings  
from the File menu, then selecting Restore Processing  
Settings. For more information, see Section 1.4.2,  
Customizing Processing and Graphic Settings (.SET).  
NOTE: All processing settings, not just calibration  
settings, are applied to the new file.  
Specifying To specify automatic calibration settings:  
1. Click the Spectrum window to activate it, then select  
the spectrum of interest.  
5
Data ExplorerSoftware Users Guide  
5-29  
               
Chapter 5 Examining Spectrum Data  
2. From the Peaks menu, select Peak Label, then set the  
Mass Label Type to Centroid.  
NOTE: For spectra containing broad peaks that have  
unresolved adducts or impurities such as proteins, you  
may obtain better results if you use apex instead of  
centroid settings.  
3. From the Process menu, select Mass Calibration, then  
select Automatic Calibration.  
The Automatic Calibration Settings dialog box is  
displayed (Figure 5-7).  
4. Click Select New File, then select a calibration reference  
file. For information on creating a reference file, see  
Section 5.3.3, Creating or Modifying a  
Calibration Reference File (.REF).  
5
5-30  
Applied Biosystems  
   
Automatic Calibration  
5. To add up to 10 reference masses to the Masses to Match  
list, do either of the following:  
Click Add All to add the first 10 reference masses  
from the reference file.  
Click Add Reference To List to individually select  
reference masses to add.  
NOTE: If the current list already contains  
10 reference masses, you must delete a mass  
before you click Add References to List.  
If you click Add Reference to List, the Select or Create  
Reference Peak Information dialog box (Figure 5-8) is  
displayed and lists all masses in the reference file. If  
you click Add All, this dialog box is not displayed.  
5
Figure 5-8 Select or Create Reference Peak  
Information Dialog Box  
Data ExplorerSoftware Users Guide  
5-31  
   
Chapter 5 Examining Spectrum Data  
6. Specify reference masses to add by doing either of the  
following:  
Click a mass, then click OK.  
Type new reference mass information in the  
Name, Theoretical m/z, Charge, and Elemental  
Composition fields, then click OK.  
NOTE: You must type a minus sign () preceding  
the charge in the Charge text box for negative  
charge states.  
NOTE: You can save changes to the Calibration  
Reference File by clicking Save. For more  
information, see Modifying a calibration reference  
fileon page 5-20.  
The Automatic Calibration Settings dialog box is  
displayed again (see Figure 5-7 on page 5-30).  
7. Repeat step 5 and step 6 to add all needed masses.  
8. To delete masses, do either of the following:  
Select a reference, then click Delete Selected  
Reference  
Click Delete All References  
9. Enter Reference Matching Criteria:  
Minimum IntensityPeaks must be above this  
intensity to be considered a match. Select the unit  
for Minimum Intensity, % Relative Intensity or  
Absolute Counts.  
5
Mass TolerancePeaks must be within this  
tolerance of the theoretical m/z to be considered a  
match. Select the unit, m/z or ppm.  
5-32  
Applied Biosystems  
   
Automatic Calibration  
10. Select the Peak Weighting Factor. If the calibration  
includes more than two points, you can apply the  
following weighting factors to fit points to the curve:  
NoneAll peaks are weighted equally  
Inverse WidthNarrower peaks are weighted  
more than broader peaks  
HeightMore intense peaks are weighted more  
than less intense peaks  
11. Enter Fit Rejection parameters:  
Minimum Peaks to MatchMinimum number of  
peaks that must meet the matching criteria for this  
calibration to be successful.  
Maximum Outlier Error (m/z or ppm)Tolerance  
within which all matched peaks must fall for this  
calibration to be successful.  
NOTE: If the software finds fewer than three  
matches, the maximum Outlier Error is not  
significant. Masses are labeled with the specified  
reference masses.  
12. Click Save Settings to save the automatic calibration  
settings (reference masses, matching criteria, and fit  
rejection parameters) as part of processing settings in  
the .DAT file.  
Saving the .SET If you are preparing Automatic Calibration settings for use by  
the Voyager Sequence Control Panel, save the .SET file by  
selecting Settings from the File menu, then select  
file  
Save Graphic/Processing Settings As.  
5
Data ExplorerSoftware Users Guide  
5-33  
   
Chapter 5 Examining Spectrum Data  
5.4.3 Automatically Calibrating  
(Mariner Data Only)  
This section includes:  
Automatically calibrating  
Applying new constants to the data file  
Calibration results  
Applying auto calibration settings to other files  
Automatically To automatically calibrate:  
calibrating  
1. Open the data file to calibrate.  
2. Click the Spectrum window.  
3. From the Process menu, select Mass Calibration, select  
Auto Calibrate State, then select On.  
NOTE: The Auto Calibrate State command setting is  
stored in a .DAT file. That is, if you turn on this  
command, then close the .DAT file, the command is on  
the next time you open the .DAT file.  
NOTE: The Auto Calibrate State command is dimmed  
unless you specified automatic calibration settings in  
the data file. See Section 5.4.2, Importing and  
Specifying Automatic Calibration Settings.  
4. Display the spectrum of interest.  
The spectrum is calibrated and displayed with an AC  
trace label if calibration is successful. Calibration  
results are displayed in the Output window.  
If calibration fails, the spectrum is displayed with an  
AC-failed trace label.  
5
5-34  
Applied Biosystems  
           
Automatic Calibration  
Applying new To save the calibration constants for each spectrum in the data  
file, select Apply Calibration from the Process menu.  
constants to the  
data file  
Calibration Automatic calibration results are displayed in the Output  
window (Figure 5-9).  
results  
Figure 5-9 Automatic Calibration Results  
For each spectrum in the data file, results include:  
Number of peaks matched  
Original and new calibration constants, calibrated mass,  
and fit errors  
Successful or failed calibration message  
Applying auto To apply auto calibration settings to other files:  
calibration  
settings to  
other files  
1. Extract the .SET information from the data file  
containing the auto calibration settings by selecting  
Settings from the File menu, then selecting Save  
Processing Settings As. For more information, see  
.DAT, .RSD, and .RCD Files.  
2. Copy the settings to the new file by selecting Settings  
from the File menu, then selecting Restore Processing  
Settings. For more information, see Section 1.4.2,  
Customizing Processing and Graphic Settings (.SET).  
5
Data ExplorerSoftware Users Guide  
5-35  
   
Chapter 5 Examining Spectrum Data  
5.5 Centroiding  
NOTE: Centroiding is not supported for Mariner DAD data.  
To display peaks as centroid traces:  
1. Click the Spectrum window to activate it.  
2. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
3. From the Process menu, select Centroiding.  
The centroid spectrum is displayed with a CT trace  
label (see Figure 5-10). The height of each vertical bar  
corresponds to the original peak area.  
Figure 5-10 Centroid Spectrum  
4. To return to the original trace, see Returning to the  
original spectrumon page 5-3.  
NOTE: If the original spectrum is displayed in vertical  
bars instead of lines, select Graphic Options from the  
Display menu, click the trace tab of interest, then select  
Lines for the Line Type in the Plot Setup section of the  
dialog box. For more information, see Section 1.5,  
Setting Graphic Options.  
5
5-36  
Applied Biosystems  
                 
Mass Deconvolution (Mariner Data Only)  
5.6 Mass Deconvolution  
(Mariner Data Only)  
NOTE: Mass deconvolution is not supported for  
Mariner DAD data.  
NOTE: The Mass Deconvolution software is an option in the  
Data Explorer software. The Multiple Charge command on  
the Process menu is dimmed if you have not purchased the  
option.  
Overview You can use the Data Explorer Mass Deconvolution features  
to generate a theoretical zero-charge spectrum that  
represents the molecular mass of a protein:  
Mass DeconvolutionUse when you have a  
spectrum with clearly resolved, multiply charged peaks.  
You specify the m/z values for the peaks to include.  
This function requires at least two adjacent peaks within  
the same charge envelope.  
Convert to Zero-Charge SpectrumUse when you  
have a spectrum with overlapping charge envelopes, or  
a noisy baseline. You specify a mass range for the  
peaks to include and a mass range for the molecular  
mass.  
Mass To use the Mass Deconvolution function:  
Deconvolution  
1. Click the Spectrum window.  
2. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
3. From the Process menu, select Multiple Charge, then  
5
select Mass Deconvolution.  
Data ExplorerSoftware Users Guide  
5-37  
                 
Chapter 5 Examining Spectrum Data  
The Multiply Charged Deconvolution dialog box  
(Figure 5-11) is displayed.  
4. In the Spectrum window, right-click-drag one multiply  
charged peak.  
5. Right-click-drag a second multiply charged peak adjacent  
to the first selected peak and in the same envelope of  
charged peaks. Mass/Charge values are entered in the  
list box.  
Figure 5-11 Multiply Charged Deconvolution  
Dialog Box  
5
5-38  
Applied Biosystems  
 
Mass Deconvolution (Mariner Data Only)  
6. Select the method to use for calculation:  
AutomaticSelects additional multiply  
charged peaks based on the selected peaks  
and performs the calculation.  
NOTE: If the trace is noisy, the software may not  
accurately select additional multiply charged  
peaks.  
ManualPerforms the calculation using only  
selected peaks. Does not select additional  
multiply charged peaks.  
7. Select Apex or Centroid mass to use for the calculation.  
8. Enter the mass/charge range to include in the calculation.  
This range is used for two purposes:  
If you selected Automatic mode, the software  
searches this range for other peaks that are in the  
same charge series as the peaks selected above.  
The software includes these other peaks in the  
calculation, to improve signal-to-noise ratio and  
the accuracy of the calculation.  
Regardless of mode, the software uses all data  
points in this range to construct the deconvoluted  
spectrum.  
9. Select or enter the mass of the adduct ion to use in the  
calculation. The mass of a proton (H+) is selected by  
default.  
10. To display both the zero-charge spectrum trace and the  
deconvolution results, select Generate Zero-Charge  
Spectrum.  
5
To display only the deconvolution results, do not select  
Generate Zero-Charge Spectrum.  
Data ExplorerSoftware Users Guide  
5-39  
Chapter 5 Examining Spectrum Data  
11. Click OK.  
The result is displayed in the Output window and the  
zero-charge spectrum is displayed with a DECONV  
trace label, if selected.  
NOTE: The numerical result displayed in the output  
window generally is more accurate than the  
computer-generated spectrum.  
12. To return to the original spectrum, click  
in the  
toolbar. The number of the original spectrum is  
displayed in the Select Spectrum dialog box.  
13. Click OK.  
Converting to To convert to a zero-charge spectrum:  
zero-charge  
1. Activate the Spectrum window.  
2. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
3. From the Process menu, select Multiple Charge, then  
select Convert to Zero-Charge Spectrum.  
The Zero-Charge Spectrum Conversion dialog box  
(Figure 5-12 on page 5-41) is displayed.  
4. Specify the Input Spectrum m/z range to include in the  
calculation by doing one of the following:  
Type in From and To values  
Right-click-drag over a region of the trace that  
includes the m/z range of interest  
5
5-40  
Applied Biosystems  
     
Mass Deconvolution (Mariner Data Only)  
5. Type values for the following masses for the generated  
zero-charge spectrum:  
CenterCenter mass  
Half WidthMass (±) from the Center mass  
to include in the spectrum  
IncrementMass increment at which to  
perform the calculation (0.1 for resolved  
isotope peaks, 1.0 for unresolved isotope  
peaks, >1.0 for noisy trace)  
Figure 5-12 Zero-Charge Spectrum Conversion  
Dialog Box  
6. Select or type the mass of the adduct ion to use in the  
calculation. The mass of a proton (H+) is selected by  
default.  
7. Click OK.  
The deconvoluted trace (with a DECONV trace label)  
replaces the original trace.  
5
8. To return to the original trace, see Returning to the  
original spectrumon page 5-3.  
Data ExplorerSoftware Users Guide  
5-41  
   
Chapter 5 Examining Spectrum Data  
5.7 Noise Filtering/Smoothing  
Description The Noise Filter/Smooth function includes four options for  
reducing noise in a spectrum trace:  
Default smoothing  
Noise-filtering  
Smoothing by the Gaussian method  
Noise removal  
Procedure To noise-filter or smooth the display:  
1. Select Duplicate Active Trace from the Display menu  
to keep the original data displayed after processing.  
2. From the Process menu, select Noise Filter/Smooth.  
The Noise Filter/Smooth dialog box (Figure 5-13) is  
displayed.  
Figure 5-13 Noise Filter/Smooth Dialog Box  
NOTE: The smoothing/filtering method selected in this  
dialog box is the method performed when you click  
in the toolbar.  
5
5-42  
Applied Biosystems  
         
Noise Filtering/Smoothing  
3. Select the method to use based on the type of data you  
are examining, then enter the associated value displayed  
for the method you select:  
Suggested  
Description  
Method  
Type of Data  
Noisy, low-resolution Default  
No associated value is displayed. Default  
data smoothing smoothing is a self-adjusting Gaussian filter  
(RSM)  
that uses the Peak Resolution (specified in the  
Peak Detection Setup dialog box) to calculate  
the optimum number of smoothing points to  
apply at every mass point. See Figure 3-3 on  
page 3-12.  
(May affect  
peak  
resolution.)  
NOTE: Default smoothing is not available for  
PSD data.  
High-resolution data, Noise  
Specify a Correlation Factor of 0 to 1.0.  
but the Noise  
Removal method  
may yield better  
results  
Filter (NF)  
Settings from 0.5 to 0.7 yield acceptable  
results for most data. A setting close to 1.0  
(May affect  
peak  
If applying the Noise Filter with a certain  
resolution.) Correlation Factor does not yield the  
necessary noise removal, return to the original  
trace (see Returning to the original spectrum”  
on page 5-3) and apply the Noise Filter again  
with a higher Correlation Factor setting.  
Applying the Noise Filter more than one time  
with the same Correlation Factor setting does  
not improve noise removal.  
Noisy, low-resolution Gaussian  
Specify a Filter Width in data points (odd  
integers only). The maximum number of  
smoothing points is 2001. Points less than 1  
Filter Width from the edge of the spectrum are  
not smoothed.  
data, chromatogram  
data, or Voyager  
PSD data  
Smooth  
(SM)  
5
(May affect  
peak  
resolution.)  
Data ExplorerSoftware Users Guide  
5-43  
     
Chapter 5 Examining Spectrum Data  
Suggested  
Type of Data  
Description  
Method  
High-resolution data  
Noise  
Removal  
(NR)  
Specify the number of standard deviations of  
noise to remove. The software automatically  
calculates the average white noise for all  
frequencies across the spectrum, then  
removes the specified number of standard  
deviations of noise.  
(Does not  
affect peak  
resolution.)  
This method slightly affects peak intensity, and  
removes peaks with a signal-to-noise ratio less  
than the specified standard deviation.  
In general:  
If you set  
Std Dev to:  
% White Noise  
removed is:  
1
2
3
68  
99  
If applying Noise Removal with a certain Std  
Dev does not yield the necessary noise  
removal, return to the original trace (see  
Returning to the original spectrumon  
page 5-3) and apply Noise Removal again with  
a higher Std Dev setting. Applying Noise  
Removal more than one time with the same  
Std Dev setting does not improve noise  
removal.  
4. Click OK.  
The trace is displayed with an RSM, NF, NR or SM  
trace label.  
5
5. To return to the original trace, see Returning to the  
original spectrumon page 5-3.  
5-44  
Applied Biosystems  
       
Adjusting the Baseline  
5.8 Adjusting the Baseline  
This section includes:  
Using Baseline Offset  
Using Baseline Correction  
Using Advanced Baseline Correction  
5.8.1 Using Baseline Offset  
Use the Baseline Offset command to offset the y-axis in a  
spectrum or to correct a sloping baseline:  
1. Activate the window in which you want to perform the  
offset.  
NOTE: You can select a Chromatogram or a Spectrum  
window. If you do not activate the correct type of  
window before performing the next step, the software  
does not select values when you click-drag on the  
trace. For example, you must select a Spectrum  
window before starting baseline offset on a spectrum.  
2. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
3. From the Process menu, select Baseline Offset.  
The Baseline Offset dialog box (Figure 5-14) is  
displayed.  
5
Data ExplorerSoftware Users Guide  
5-45  
     
Chapter 5 Examining Spectrum Data  
Figure 5-14 Baseline Offset Dialog Box  
4. Right-click-drag the left baseline to offset. The selected  
value is displayed in the Left Baseline field.  
5. Right-click-drag the right baseline to offset. The selected  
value is displayed in the Right Baseline field.  
6. To limit the baseline offset to the area between the two  
selected points, select Only Apply from L to R Baseline  
Midpoint.  
To perform the baseline offset on the entire x-axis,  
deselect Only Apply from L to R Baseline Midpoint.  
7. Click OK.  
The offset baseline trace is displayed with a BO trace  
label.  
8. To return to the original trace, see Returning to the  
original spectrumon page 5-3.  
5
5-46  
Applied Biosystems  
   
Adjusting the Baseline  
5.8.2 Using Baseline Correction  
Description The Baseline Correction function corrects for a curved  
baseline, including a DC-offset baseline, by eliminating broad  
artifacts from the data set.  
When to use Baseline-correct if you are analyzing data:  
With a baseline that is not flat, and you are using the  
%Base Peak Intensity parameter (intensity-based  
thresholding) to screen out noise peaks. For best results,  
apply Baseline Correction, then re-detect peaks.  
If you use area-based thresholding (%Max Peak Area)  
during peak detection, Baseline Correction is not typically  
needed. %Max Peak Area compensates for a rising or  
falling baseline.  
With a baseline that is not at 0. Some measurements (for  
example, Centroid% peak detection value, or the Peak  
Height % in the Resolution calculator), are derived from a  
peak height measured from 0.  
NOTE: If you are analyzing Mariner data, baseline  
correction is typically not needed. If you baseline-correct  
Mariner data, note that due to the shorter flight times and  
fewer data points associated with Mariner data, baseline  
correction may affect peak shape, which in turn affects  
mass accuracy.  
Correcting the To correct the baseline:  
baseline  
1. Display the spectrum of interest.  
2. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
3. From the Process menu, select Baseline Correction.  
The baseline is adjusted, and the trace is displayed  
with a BC trace label.  
5
Returning to the To return to the original trace, see Returning to the original  
spectrumon page 5-3.  
original spectrum  
Data ExplorerSoftware Users Guide  
5-47  
           
Chapter 5 Examining Spectrum Data  
5.8.3 Using Advanced Baseline Correction  
This section includes:  
Description  
When to use  
Correcting the baseline  
General guidelines for setting parameters  
Troubleshooting  
Returning to the original spectrum  
Description The Advanced Baseline Correction feature corrects the  
baseline by:  
Iteratively estimating baseline amplitude at regularly  
spaced intervals throughout the spectrum  
Smoothly connecting the calculated baseline points  
Removing the calculated baseline from the spectrum  
x
x
x
x
x
x
x
x
x
x
x
x
x
x
5
Figure 5-15 Advanced Baseline Correction  
5-48  
Applied Biosystems  
     
Adjusting the Baseline  
When to use Use advanced baseline correction if you are analyzing data  
with an offset in the spectrum, particularly data with a strong  
sloping baseline at low mass.  
NOTE: Because this function is iterative, it may take  
several seconds to complete, and typically takes longer for  
narrower peaks.  
Correcting the To correct the baseline:  
baseline  
1. Select Duplicate Active Trace from the Display menu  
to keep the original data displayed after processing.  
2. From the Process menu, select Advanced Baseline  
Correction.  
The Advanced Baseline Correction dialog box  
(Figure 5-16) is displayed.  
Figure 5-16 Advanced Baseline Correction Dialog Box  
5
Data ExplorerSoftware Users Guide  
5-49  
 
Chapter 5 Examining Spectrum Data  
3. Enter parameters as described below. These parameters  
interact with each other and require experimentation to  
determine the optimum settings for your data. Refer to  
General guidelines for setting parameterson page 5-54  
for more information.  
Parameter  
Description/Specifies  
Peak Width  
(at half height)  
Value the software uses to estimate the baseline  
amplitude at regularly spaced points in the spectrum. The  
number of regularly spaced points used is derived using  
the Peak Width parameter and the Flexibility parameter.  
The illustration below shows how using a higher number  
of points estimate the baseline (accomplished by using  
higher Peak Width and Flexibility settings) can affect the  
shape and slope of the calculated baseline.  
Lower number of points  
x
x
x
x
x
x
Higher number of points  
x
x
x
x
x
x
x
x
x
x
x
x
(continued)  
5
5-50  
Applied Biosystems  
Adjusting the Baseline  
Parameter  
Peak Width  
Description/Specifies  
Set Peak Width according to the data you are correcting:  
(at half height)  
For best results, set to the peak width at half height  
of the narrowest peak. However, smaller peak width  
values increase processing time.  
(continued)  
If peak width varies across the spectrum, set to the  
average peak width.  
If peaks are narrow relative to the baseline region  
you are correcting, set a larger value (for example,  
10 times the width of widest peak) to increase  
processing speed.  
If the baseline changes sharply across the spectrum,  
set a smaller value closer to the narrowest peak  
width.  
If the baseline is broad or gently sloping, set a larger  
value, for example 10 to 20 times the width of the  
actual peak width.  
To determine the number of points across a peak, change  
the trace display from Line to Vertical Bars (select  
Graphic Options from the Display menu, click Graph #  
Setup and change Line Type). Each vertical bar  
represents one data point. For more information, see  
Section 1.4, Customizing the Data Explorer Window.  
(continued)  
5
Data ExplorerSoftware Users Guide  
5-51  
 
Chapter 5 Examining Spectrum Data  
Parameter  
Description/Specifies  
Flexibility  
With the Peak Width parameter, determines the number  
of points used to estimate the baseline amplitude at  
regularly spaced points in the spectrum, but is not directly  
proportional to the number of points used. Valid entries  
are 0 to 1.0.  
The default value of 0.5 works best for most applications.  
A value closer to 0 reduces flexibility and provides a  
smoother more generalized baseline correction (see  
below).  
A value closer to 1 provides more localized baseline  
correction by using a larger number of points to estimate  
the baseline (see below).  
If the baseline rises adjacent to peaks after  
baseline-correcting, decrease the Flexibility value and  
baseline-correct again.  
Flexibility=0  
Flexibility=1  
(continued)  
5
5-52  
Applied Biosystems  
Adjusting the Baseline  
Parameter  
Degree  
Description/Specifies  
Determines how closely the calculated baseline fits the  
data. Valid entries are 0.0 to 1.0.  
A value closer to 1 fits the baseline more closely to the  
data, and corrects the midpoint of the noise signal to  
approximately 0 intensity.  
A value closer to 0 fits the baseline less closely to the  
data, and corrects the midpoint of the noise signal to a  
value greater than 0 intensity (see below).  
For protein and tryptic digest peptide spectra, use 0 to  
0.5. If the baseline after correction is too high, increase  
the Degree value (see below).  
Degree=1  
Midpoint of noise signal  
corrected to 0  
Degree=0  
Midpoint of noise signal  
corrected above 0  
A lower Degree setting allows the baseline correction to  
occur more quickly, and often provides better correction  
than a higher setting.  
NOTE: Flexibility and Degree parameters may require  
some testing to determine the optimum settings for your  
data.  
5
4. Click OK.  
The baseline is adjusted, and the trace is displayed with a  
AdvBC trace label.  
Data ExplorerSoftware Users Guide  
5-53  
 
Chapter 5 Examining Spectrum Data  
General Refer to the following table to determine how to set Advance  
Baseline Correction parameters and obtain the desired  
baseline correction.  
guidelines for  
setting  
parameters  
Condition  
Set Parameters To  
Baseline follows a rising peak cluster  
Peak WidthAverage peak width  
Flexibility0.5  
Degree0.5  
Baseline is gently curving  
Peak Width10 to 20 times peak width  
Flexibility0  
Degree0.5 (decrease further to raise  
the baseline)  
Baseline rise should be ignored and  
treated as signal  
Peak WidthHalf of the width of the  
cluster to correct (in the illustration  
shown, if the width of the cluster is 200,  
set Peak Width to 100)  
Peak cluster  
Flexibility0  
Degree0.5 (decrease further to raise  
the baseline)  
5
5-54  
Applied Biosystems  
 
Adjusting the Baseline  
Troubleshooting If the baseline rises preceding and following a peak after the  
correction (a humpunder the peak), adjust the following  
parameters in the order listed:  
Decrease Flexibility  
Decrease Degree  
Decrease Peak Width  
NOTE: Lower Peak Width values increase the time needed  
for processing.  
Returning to the To return to the original trace, see Returning to the original  
spectrumon page 5-3.  
original spectrum  
5
Data ExplorerSoftware Users Guide  
5-55  
Chapter 5 Examining Spectrum Data  
5.9 Truncating a Spectrum  
Description The Truncate function removes data points from a trace  
outside a selected region.  
Truncating spectra is useful to:  
Mariner dataRemove noise at the low end of a  
spectrum before generating a result file  
Voyager dataEliminate the Low Mass Gate spike and  
background in the low-mass range  
Truncating To truncate spectra:  
1. Display the spectrum to truncate.  
2. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
3. From the Process menu, select Truncate Spectrum.  
The Truncate Spectrum dialog box (Figure 5-17) is  
displayed.  
Figure 5-17 Truncate Spectrum Dialog Box  
4. Type the starting and ending m/z values for the range of  
the spectrum to retain, or right-click-drag over the range in  
the spectrum.  
5
5. Click OK.  
5-56  
Applied Biosystems  
       
Truncating a Spectrum  
The data in the spectrum is truncated to the selected  
displayed on the axis of the truncated trace is the range  
of the original data file, and may be wider than the  
range of the truncated spectrum.  
6. To return to the original trace, see Returning to the  
original spectrumon page 5-3.  
Example Figure 5-18 and Figure 5-19 illustrate the effects of  
truncating a Voyager spectrum and eliminating the Low  
Mass Gate peak.  
Figure 5-18 includes a Low Mass Gate spike. Because the  
Low Mass Gate spike is the most intense peak in the  
spectrum, it is identified as the base peak, and:  
All other peaks in the spectrum are scaled as a  
percentage of the base peak.  
A default peak detection threshold, %Base Peak  
Intensity, is set to 1 percent of the base peak. One  
percent of an intense peak yields a high default threshold,  
and many peaks in the spectrum are not detected.  
Low Mass Gate spike  
Figure 5-18 Complete Spectrum with Low Mass Gate  
Spike  
Figure 5-19 shows the spectrum truncated to eliminate the  
Low Mass Gate spike. The peak at 3772 Da is identified as the  
base peak, the %Base Peak Intensity is set to a lower value,  
and more peaks in the spectrum are detected.  
5
Data ExplorerSoftware Users Guide  
5-57  
     
Chapter 5 Examining Spectrum Data  
Figure 5-19 Truncated SpectrumLow Mass Gate Spike Eliminated  
5
5-58  
Applied Biosystems  
 
Converting to a Singly Charged Spectrum (Mariner Data Only)  
5.10 Converting to a Singly Charged  
Spectrum (Mariner Data Only)  
NOTE: Single-charge conversion is not supported for  
Mariner DAD data.  
Description The Single-Charge Conversion function generates a  
theoretical, centroided, singly charged spectrum. This function  
uses isotopic spacing in detected spectral peaks to generate  
the theoretical spectrum.  
For example, if you use this function on a spectrum that  
includes +3 peaks, the software multiplies each mass by 3 and  
removes the mass of two extra protons. The resulting  
spectrum contains theoretical +1 species.  
If more than one charge state is present, each charge state is  
converted appropriately, and the total range of the converted  
spectrum is approximately 1.2 times the m/z of the highest m/z  
present.  
Requirements Use this function only on isotopically resolved data that is  
labeled with the correct charge state. Peaks with low  
signal-to-noise ratios may be labeled with incorrect charge  
states. Set peak detection thresholds to disregard these peaks  
before converting the spectrum.  
If you use this function on a spectrum that is not correctly  
labeled, the resulting spectrum may contain charge states  
other than 1.  
Converting To convert a multiply charged spectrum to a singly charged  
spectrum:  
1. Display the spectrum trace of interest.  
2. Examine the spectrum. Make sure peaks are labeled with  
the correct charge state, and that no noise peaks are  
labeled with a charge state. Adjust peak detection  
parameters as needed.  
5
Data ExplorerSoftware Users Guide  
5-59  
         
Chapter 5 Examining Spectrum Data  
3. Select Duplicate Active Trace from the Display menu to  
keep the original data displayed after processing.  
4. From the Process menu, select Single-Charge  
Conversion.  
The Single-Charge Spectrum Conversion dialog box  
(Figure 5-20) is displayed.  
Figure 5-20 Single-Charge Spectrum Conversion  
Dialog Box  
5. In the Adduct text box, type or select the adduct that is the  
charge-carrying species in the spectrum you are  
examining.  
6. Select the polarity for the converted spectrum. You must  
select the same charge as the spectrum you are  
evaluating.  
CAUTION  
If you do not select the same charge as the spectrum  
you are evaluating, an incorrect mass is reported.  
7. Click OK.  
The converted spectrum is displayed with an SC trace  
label. The height of each vertical bar corresponds to the  
original peak areas.  
5
8. To return to the original trace, see Returning to the  
original spectrumon page 5-3.  
5-60  
Applied Biosystems  
   
Converting to a Singly Charged Spectrum (Mariner Data Only)  
Example Figure 5-21 and Figure 5-22 illustrate the effects of  
single-charge conversion.  
Before conversion (Figure 5-21), the spectrum includes +2  
and +3 charged species of neurotensin.  
Neurotensin  
multiply charged  
species  
Figure 5-21 Spectrum Before Single-Charge Conversion  
After conversion (Figure 5-22), the +2 and +3 charged  
species are converted to the +1 species of neurotensin.  
Neurotensin  
singly charged  
species  
Figure 5-22 Spectrum After Single-Charge Conversion  
5
Data ExplorerSoftware Users Guide  
5-61  
     
Chapter 5 Examining Spectrum Data  
NOTE: Charge states other than 0 or 1 in the converted  
trace indicate that a peak in the original spectrum is labeled  
with an incorrect charge state. Set peak detection  
thresholds to disregard these peaks and convert the  
spectrum again.  
CAUTION  
A zero value in the Spec Peak list does not indicate a  
charge state of zero. It indicates that the software could not  
determine the charge state.  
5.11 AutoSaturation Correction  
(Mariner Data Only)  
NOTE: Autosaturation correction is not supported for  
Mariner .RST or DAD data.  
Function The AutoSaturation Correction function mathematically  
corrects for signal saturation of the Mariner detector system to  
provide optimum mass accuracy.  
The AutoSaturation Correction feature is turned on by default,  
and automatically corrects the data when you open a Mariner  
data file in Data Explorer. Leave this feature turned on for  
normal operation.  
Hardware The AutoSaturation Correction function requires special  
hardware. If the system used to acquire the data does not  
requirements  
5
include the required hardware, the AutoSaturation Correction  
does not completely correct for saturation of the detection  
system, even if it is enabled.  
5-62  
Applied Biosystems  
             
AutoSaturation Correction (Mariner Data Only)  
Effect on Mariner The AutoSaturation Correction feature is not applied to .RST  
files saved from the Mariner Instrument Control Panel, even  
.RST files  
when Saturation Correction is turned on. Saturation requires  
information about the pulser frequency used to acquire the  
data, and this information is not stored in .RST files saved  
from the Instrument Control Panel.  
Trace label When AutoSaturation Correction is turned on, spectra in the  
data file are displayed with an ASC (AutoSaturation  
Correction) trace label.  
5
Data ExplorerSoftware Users Guide  
5-63  
 
Chapter 5 Examining Spectrum Data  
5.12 Adding and Subtracting Raw or  
Processed Spectra from the  
Same or Different Data Files  
(Dual Spectral Trace Arithmetic)  
The Dual Spectral Trace Arithmetic function lets you add two  
spectra together, or subtract one spectrum from another.  
Spectra can be raw or processed, and can be from the same  
To use the Dual Spectral Trace Arithmetic function:  
1. To perform this function on spectra from different data  
files, copy a spectrum trace from one data file to  
another. See Section 2.5.2, Copying Traces from  
Multiple Data Files to a Window, for information on  
copying traces.  
2. Activate the Spectrum window that contains the two  
traces of interest.  
3. Process the traces as needed.  
4. Click the first trace in the Spectrum window to make it the  
active trace.  
NOTE: The Dual Spectral Trace Arithmetic command is  
dimmed unless the first trace in the Spectrum window is  
the active trace.  
5. From the Process menu, select Dual Spectral Trace  
Arithmetic.  
The Dual Spectral Trace Arithmetic dialog box is  
displayed (Figure 5-23).  
5
5-64  
Applied Biosystems  
         
Adding and Subtracting Raw or Processed Spectra from the Same or Different Data Files (Dual  
Figure 5-23 Dual Spectral Trace Arithmetic Dialog Box  
6. Set the Mass Tolerance within which data points from the  
different traces will be considered as the same mass.  
7. Select Add or Subtract for Operation.  
8. Select Add New Trace or Replace Active Trace for the  
result trace.  
9. Click OK.  
Absolute intensities of the two traces are added or  
subtracted, and the result trace is displayed with ADD  
or SUB in the trace header.  
5
Data ExplorerSoftware Users Guide  
5-65  
 
Chapter 5 Examining Spectrum Data  
5
5-66  
Applied Biosystems  
Chapter  
6
6 Using Tools and  
Applications  
This chapter contains the following sections:  
6.1  
Using the  
Elemental Composition Calculator .......... 6-2  
6.2  
6.3  
Using the Isotope Calculator ................ 6-13  
Using the  
Mass Resolution Calculator .................. 6-20  
6.4  
6.5  
6.6  
6.7  
Using the  
Signal-to-Noise Ratio Calculator........... 6-23  
Using the  
Ion Fragmentation Calculator ............... 6-25  
Using the  
Elemental Targeting Application............ 6-31  
Using the Macro Recorder.................... 6-34  
Data ExplorerSoftware Users Guide  
6-1  
   
Chapter 6 Using Tools and Applications  
6.1 Using the Elemental  
Composition Calculator  
6
This section includes:  
Determining elemental composition  
Setting limits  
6.1.1 Determining Elemental Composition  
This section includes:  
Description  
Determining identity of fragment masses  
Procedure  
Calculating  
Results  
If no results are displayed  
Displaying the theoretical isotope distribution  
Description The Elemental Composition calculator determines possible  
elemental or amino acid compositions for a given mass. The  
application then generates a theoretical isotope pattern using  
the Mass Resolution specified in Basic Peak Detection  
settings, compares each observed mass and isotope pattern  
to the theoretical mass and isotope pattern for each possible  
composition, and reports an isotope match score that reflects  
how closely they match.  
Determining You can also use the Elemental Composition calculator to  
determine the identity of fragment masses. To do so, set Peak  
Labels to display Mass Difference From Selected Peak, then  
specify the mass of the intact molecule. For more information,  
identity of  
fragment masses  
see Setting spectrum labelson page 3-56.  
6-2  
Applied Biosystems  
               
Using the Elemental Composition Calculator  
Procedure To determine elemental composition:  
1. Display the spectrum containing the peak of interest.  
6
2. Click the Spectrum window to activate it.  
3. From the Applications menu, select Elemental  
Composition.  
The Elemental Composition Calculator dialog box  
(Figure 6-1) is displayed.  
Figure 6-1 Elemental Composition Calculator  
Dialog Box  
Data ExplorerSoftware Users Guide  
6-3  
   
Chapter 6 Using Tools and Applications  
4. Enter m/z values in the m/z ratio list by doing any of the  
following:  
6
Right-click-drag over a peak in the spectrum to  
add the m/z and the associated charge state.  
Double-click the line to display the Elemental  
Target Mass dialog box, type the m/z and the  
associated charge state, then click OK.  
Click  
to add a line to the list, then  
double-click the line to display the Elemental  
Target Mass dialog box, type the m/z and the  
associated charge state, then click OK.  
5. Specify the Tolerance to use. The Tolerance value entered  
is directly used as the window for elemental comparisons,  
and is also multiplied by 2 and used as the window for  
isotope peak comparison (because related isotope peaks  
will usually have lower mass accuracy than monoisotopic  
peak). Setting Tolerance too low can result in a falsely low  
6. Specify the Mass Type to calculate.  
7. If the displayed Mass Peak Resolution is not appropriate  
for this calculation, change the setting in Peak Detection.  
For more information, see Section 3.2.4, Peak Detection  
Parameter Descriptions.  
8. Specify the Result Type (Elemental, Amino Acid,  
Carbohydrate, DNA, or RNA).  
9. Specify the Max Number of Results for each m/z entered  
in step 4.  
NOTE: Max Number of Results is per m/z entered, not  
total number of results generated for the calculation.  
10. Set limits as described in:  
Setting limits for existing elementson  
page 6-7  
6-4  
Applied Biosystems  
 
Adding new elements and setting limitson  
page 6-9  
6
Setting limits for other result typeson  
page 6-12  
11. Click More Parameters, then enter the Minimum and  
Maximum Double Bond Equivalents to include in the  
calculation.  
Electron state for 12. Set the Electron State to calculate. If you are  
calculating composition for:  
intact molecules or  
fragment ions  
Intact moleculesUse the default of Even  
Only.  
Fragment ionsSet to Both Odd and Even  
if you are analyzing mass differences or  
absolute masses.  
NOTE: You may be able to refine results for fragment  
ions by setting Electron State to Odd Only.  
13. Click OK.  
Calculating 14. Click Calculate.  
To cancel a lengthy calculation, click Cancel Calc.  
Results The results of the calculation are displayed in the Elemental  
Analysis tab of the Output window (Figure 6-2).  
Figure 6-2 Elemental Composition Calculation Results in Output Window  
Data ExplorerSoftware Users Guide  
6-5  
     
Chapter 6 Using Tools and Applications  
Hint: You can sort the results in a column by clicking the  
column header.  
6
Results include:  
IndexSequential number assigned to each result.  
Input m/zEntered m/z for each composition  
calculation.  
Calculated MassCalculated mass for the entered m/z  
and charge state for each theoretical composition.  
mDa Error and ppm ErrorError for each calculation.  
Double Bond Equivalents (DBE)Value that  
corresponds to the number of double bonds or rings that  
the valence-bond theory requires to construct a molecule  
from each calculated composition. If the target mass is a  
protonated molecular ion, (M+H)+, the DBE value  
includes 0.5 and corresponds to an even electron state.  
To convert the reported DBE value to the actual number  
of double bonds or rings, add 0.5. For example,  
H2C=NH2+ has an (M+H)+ total mass of 30.0344 and a  
reported double bond equivalent of 0.5, with an actual  
number of double bonds of 1.  
FormulaElemental composition for each mass.  
Isotope Match ScoreNumber between 0 and 1.0 that  
reflects how well the observed peak matches the  
theoretical formula, based on mass and isotope pattern.  
A higher number represents a better match.  
NOTE: An Isotope Match Score of 0.00000 is always  
reported for fragment ion calculations.  
If no results are If no results are displayed, it may indicate that the specified  
Tolerance is too limiting and no matches were found.  
displayed  
6-6  
Applied Biosystems  
         
Using the Elemental Composition Calculator  
Displaying the To display the theoretical isotope distribution for a calculated  
formula, double-click the corresponding line in the Elemental  
Analysis tab of the Output window.  
theoretical  
isotope  
distribution  
6
The theoretical isotope trace is displayed in the Spectrum  
window with an ISO trace label and the elemental formula.  
To return to the original trace, see Returning to the original  
spectrumon page 5-3.  
6.1.2 Setting Limits  
This section includes:  
Setting limits for existing elements  
Adding new elements and setting limits  
Setting limits for other result types  
Setting limits for To set limits for existing elements:  
existing elements  
1. To set limits for Elemental results, click Element Limits  
in the Elemental Composition Calculator dialog box.  
The Limits dialog box is displayed (Figure 6-3).  
Figure 6-3 Element Limits Dialog Box  
Data ExplorerSoftware Users Guide  
6-7  
           
Chapter 6 Using Tools and Applications  
2. To change the limits for an element, double-click an  
element to display the Isotope dialog box (Figure 6-4).  
6
Figure 6-4 Isotope Dialog Box  
NOTE: Ignore the column of check boxes to the left of  
the Isotope column if it is displayed.  
3.  
Change the Minimum and Maximum number of  
occurrences for the element as needed.  
NOTE: The software ignores changes you make to the  
individual isotope minimum and maximum values.  
4.  
5.  
Click OK.  
Change limits for other elements as needed, then click  
OK to return to the Elemental Composition Calculator.  
6-8  
Applied Biosystems  
   
Using the Elemental Composition Calculator  
Adding new To add new elements and set limits:  
elements and  
setting limits  
1. To add new elements and set limits for Elemental  
results, click Element Limits in the Elemental  
Composition Calculator dialog box.  
6
The Limits dialog box is displayed (Figure 6-5).  
Figure 6-5 Element Limits Dialog Box  
2. Click  
. The Periodic Table (Figure 6-6) is displayed.  
Data ExplorerSoftware Users Guide  
6-9  
         
Chapter 6 Using Tools and Applications  
6
Figure 6-6 Periodic Table  
3. Click an element to select it and to display the Isotope  
dialog box (Figure 6-7).  
6-10  
Applied Biosystems  
   
Using the Elemental Composition Calculator  
6
Figure 6-7 Isotope Dialog Box  
NOTE: Ignore the column of check boxes to the left of  
the Isotope column if it is displayed.  
4. Change the Minimum and Maximum number of  
occurrences for the first isotope of the element as needed.  
NOTE: The software ignores changes you make to the  
individual isotope minimum and maximum values.  
5. Click OK two times to return to the Element Limits dialog  
box.  
6. Repeat step 2 through step 5 to add other elements as  
needed, then click OK to return to the Elemental  
Composition Calculator.  
Data ExplorerSoftware Users Guide  
6-11  
     
Chapter 6 Using Tools and Applications  
Setting limits for To set limits for amino acid, DNA, RNA, or carbohydrate result  
types:  
other result types  
6
1. Click the limits button displayed for the selected result  
type in the Elemental Composition Calculator dialog box  
(see Figure 6-1 on page 6-3).  
The Limits dialog box for the selected Result Type is  
displayed. Figure 6-8 shows the Amino Acid Limits  
dialog box.  
Figure 6-8 Amino Acid Limits Dialog Box  
2. Change the Min Number and Max Number for the  
compound for which you are setting limits, then click OK.  
3. Change limits for other entries as needed, then click OK  
to return to the Elemental Composition Calculator.  
6-12  
Applied Biosystems  
     
Using the Isotope Calculator  
6.2 Using the Isotope Calculator  
6
Description Use the Isotope calculator to generate a theoretical isotope  
distribution. You can compare or overlay the theoretical  
distribution with your observed distribution.  
Using the Isotope To use the Isotope calculator:  
Calculator  
1. Display the spectrum containing the observed isotope  
distribution.  
2. Click the Spectrum window to activate it.  
3. From the Applications menu, select Isotope Calculator.  
The Isotope Calculator dialog box (Figure 6-9) is  
displayed.  
Figure 6-9 Isotope Calculator Dialog Box  
Data ExplorerSoftware Users Guide  
6-13  
       
Chapter 6 Using Tools and Applications  
4. Select the Formula or Sequence for the type of isotope to  
calculate.  
6
5. Select a formula from the list, or type in a new formula.  
Valid entries for each formula type are:  
Formula Type  
Valid Entries  
Elemental  
Any element from the Periodic  
Table. The first letter of two-letter  
elemental symbols must be  
capitalized. Spaces do not matter.  
Amino Acid  
Sequence  
One-letter or three-letter amino acid  
codes. The first letter of amino acid  
codes must be capitalized. Spaces  
do not matter.  
NOTE: Do not include elemental  
codes when you type the sequence  
(for example, to specify water of  
hydration). Invalid results may be  
generated.  
DNA  
ACGT  
Sequence  
RNA  
ACGU  
Sequence  
Carbohydrates Any entry listed for carbohydrates  
from Elemental Composition Limits.  
See Setting limits for other result  
typeson page 6-12.  
Separate entries with spaces.  
6. Select the Plus H2O check box if you want to calculate  
the isotopes for the complete molecule (residue plus  
water of hydration).  
6-14  
Applied Biosystems  
   
Using the Isotope Calculator  
7. Specify the Add/Subtract Group option. To enable or  
disable the option, select or deselect the Group Type  
check box.  
6
8. In the Add/Subtract Group section, select:  
Element or group to add or subtract from the  
formula before calculating the isotope (does not  
apply if Add/Subtract Group is disabled).  
Number of elements or groups to add or remove  
and the charge state to divide by.  
NOTE: When Add/Subtract is enabled, the Group  
Count/Charge field determines the number of  
groups to add or subtract, and the charge state to  
calculate. When Add/Subtract is disabled, the  
Charge field determines the charge state to  
calculate.  
Whether to add or subtract the specified number  
of elements or groups.  
NOTE: If you specify Subtract, the group to  
subtract must be present in the formula you  
specify in step 5.  
For more information, see Add and subtract examples”  
on page 6-16.  
9. Select the Resolution option:  
DaResolves peaks that are separated by the  
number of daltons you enter.  
Resolving PowerResolves peaks using the  
resolving power (M/M) you enter.  
PPMResolves peaks within the number of PPM  
(parts per million) you enter.  
Data ExplorerSoftware Users Guide  
6-15  
 
Chapter 6 Using Tools and Applications  
10. Select the calculation mode:  
6
FWHMResolves peaks using the full peak width  
at peak half height.  
10% ValleyResolves peaks to a 10 percent  
valley.  
11. Set the Threshold %. Signal intensity below this percent  
intensity is not included in the calculation or display.  
NOTE: Setting Threshold % too high distorts the  
isotope pattern.  
12. Click Calculate.  
The theoretical isotope distribution is displayed in the  
Spectrum window with an ISO trace label and the  
elemental formula, or the amino acid, DNA, or RNA  
sequence.  
The result is displayed in the Result tab in the Output  
window. For more information, see Evaluating traces”  
on page 6-18.  
Add and subtract The following examples illustrate how the software adds and  
subtracts groups to or from an elemental formula, and how it  
applies the charge state field.  
examples  
To calculate a theoretical isotope distribution for doubly  
sodiated β-cyclodextrin with a +2 charge state (m/z 545)  
The software performs  
With these parameters specified  
this calculation  
Formula*C42H60O30  
Group typeNa  
Group Count/Charge2  
Add  
(C42H60O30 + 2Na)/2  
6-16  
Applied Biosystems  
     
Using the Isotope Calculator  
To calculate a theoretical isotope distribution for doubly  
sodiated and deprotonated β-cyclodextrin with  
1 charge state (m/z 1089)  
6
The software performs  
With these parameters specified  
this calculation  
Formula*C42H60O30Na2  
Group typeH  
(C42H60O30Na2 H)  
Group Count/Charge1  
Subtract  
To calculate a theoretical isotope distribution for  
β-cyclodextrin with +2 charge state (m/z 522)  
The software performs  
With these parameters specified  
this calculation  
Formula*C42H60O30  
Group typedeselected  
Charge2  
C42H60O30/2  
Add  
*Formula entry is case sensitive  
Data ExplorerSoftware Users Guide  
6-17  
Chapter 6 Using Tools and Applications  
Evaluating traces The theoretical isotope distribution is displayed in the  
Spectrum window with an ISO trace label (Figure 6-10).  
6
Figure 6-10 Isotope Trace  
If you have the Replace mode set to Add in the Display Trace  
dialog box, a new trace is added. If you have the Replace  
trace. For more information on Replace mode, see Setting  
the Replace modeon page 2-17.  
To evaluate the trace, you can:  
Overlay theoretical and observed isotope patterns. See  
Section 2.4.8, Overlaying Traces, for information.  
Add new traces and calculate the isotope multiple times  
using different settings, then compare or overlay the  
different calculations.  
Returning to the To return to the original trace, see Returning to the original  
spectrumon page 5-3.  
original spectrum  
6-18  
Applied Biosystems  
         
Using the Isotope Calculator  
Results The results of the calculation are displayed in the Result tab of  
Output window (Figure 6-11).  
6
Figure 6-11 Isotope Calculation Results  
in Output Window  
If results are not If results are not calculated, you may have tried to remove a  
group that is not present in the formula.  
calculated  
Data ExplorerSoftware Users Guide  
6-19  
     
Chapter 6 Using Tools and Applications  
6.3 Using the  
6
Mass Resolution Calculator  
Calculating mass To calculate mass resolution:  
resolution  
Spectrum window is active.  
2. If you are examining Voyager data, baseline correct the  
spectrum. The Peak Height used by the Resolution  
calculator is calculated from 0. For information, see  
Section 5.8.2, Using Baseline Correction.  
NOTE: Baselines in Mariner data are typically at 0.  
Baseline correction is not necessary.  
3. From the Tools menu, select Resolution Calculator.  
NOTE: If the Spectrum window is not active, Resolution  
Calculator does not appear on the Tools menu.  
4. In the Resolution Calculator dialog box (Figure 6-12), set  
the percentage of peak height at which to calculate  
resolution. The default is 50%, which calculates the  
resolution at the full width/half maximum (FWHM) of a  
peak.  
NOTE: Measuring resolution at the full width/half  
maximum is the industry standard. Use 50% Peak  
Height for most applications. If a peak is not resolved at  
the Peak Height % you set, an error message is  
displayed in the Result tab.  
6-20  
Applied Biosystems  
               
Using the Mass Resolution Calculator  
6
FWHM=50% peak height  
Figure 6-12 Mass Resolution Calculator  
5. Specify the peak for which to calculate resolution by doing  
one of the following:  
Type in a Mass/Charge value.  
In the Spectrum window, right-click-drag over  
the peak for which you want to calculate  
resolution. The mass is automatically entered  
in the Resolution Calculator dialog box.  
NOTE: If you right-click-drag over more than one  
peak, the mass of the highest peak is used.  
6. Click OK.  
Data ExplorerSoftware Users Guide  
6-21  
 
Chapter 6 Using Tools and Applications  
The result is displayed in the Output window  
(Figure 6-13).  
6
Resolution  
calculator  
results  
Figure 6-13 Resolution Calculator Results  
6-22  
Applied Biosystems  
   
Using the Signal-to-Noise Ratio Calculator  
6.4 Using the Signal-to-Noise Ratio  
Calculator  
6
Description A signal-to-noise ratio is typically used to describe how well a  
peak of interest in a spectrum or chromatogram is  
distinguished from background noise.  
The Data Explorer software provides two signal-to-noise ratio  
methods that calculate signal-to-RMS noise (electronic and  
chemical):  
AutoYou specify the peak for calculation and the  
software automatically calculates the average noise  
across the spectrum.  
ManualYou specify the peak and the baseline region  
for calculation. For accurate results, this method requires  
a flat (non-rising) baseline that does not include peaks.  
The manual method is useful when evaluating a narrow  
region around a peak to determine the relative  
significance of a peak.  
Calculating To calculate a signal-to-noise ratio:  
signal-to-noise  
ratio  
1. Click the Spectrum window to activate it.  
2. From the Tools menu, select S/N Calculator.  
The Signal-to-Noise Calculator dialog box (Figure 6-14)  
is displayed.  
Figure 6-14 Signal-to-Noise Dialog Box  
Data ExplorerSoftware Users Guide  
6-23  
             
Chapter 6 Using Tools and Applications  
3. Select the method to use, then right-click-drag on peaks  
in the trace to enter the associated values displayed for  
the method you select:  
6
AutoRight-click-drag across the apex of the  
peak for signal-to-noise calculation. The  
software automatically calculates the average  
noise across the entire spectrum.  
The noise calculation is not affected by the  
presence of peaks or a poor baseline.  
ManualRight-click-drag across the apex of  
the peak for signal-to-noise calculation.  
Right-click-drag across a flat region of the  
baseline to use to determine the RMS noise  
for the calculation.  
NOTE: If you enter a baseline value instead of  
right-click-dragging, the software uses an RMS  
value of 1.000 to perform the signal-to-noise  
calculation.  
4. Click OK. The result is displayed in the Output window.  
6-24  
Applied Biosystems  
 
Using the Ion Fragmentation Calculator  
6.5 Using the  
6
Ion Fragmentation Calculator  
Description The Ion Fragmentation calculator generates a list of possible  
fragment masses for a peptide sequence you enter. It can  
calculate the masses for:  
Multiply charged fragments you may see in Mariner  
data.  
PSD fragments you may see in Voyager data.  
If the calculated fragments are present in the current data file,  
you can label fragments.  
The calculator includes a list of defined amino acids and  
residues that you can add to as needed.  
Using the Ion To use the Ion Fragmentation Calculator:  
Fragmentation  
calculator  
1. Display the spectrum of interest.  
2. Click the Spectrum window to activate it.  
3. From the Applications menu, select Ion Fragmentation  
Calculator.  
The Ion Fragmentation Calculator dialog box  
(Figure 6-15) is displayed.  
Data ExplorerSoftware Users Guide  
6-25  
           
Chapter 6 Using Tools and Applications  
6
Figure 6-15 Ion Fragmentation Calculator Dialog Box  
4. Type or copy the amino acid or residue sequence of  
interest. Use single-letter codes.  
Sequence codes are case-sensitive. Click the  
User-Defined Amino Acids button to display the list of  
allowed residues and their associated codes.  
5. Select the N-terminus and C-terminus options for the  
sequence.  
6. Specify the cysteine modification, and specify the mass  
type (average or monoisotopic) for the calculated  
fragment ion masses.  
6-26  
Applied Biosystems  
   
Using the Ion Fragmentation Calculator  
Setting Options 7. Click Options.  
6
The Ion Fragmentation Options dialog box (Figure 6-16)  
is displayed.  
Figure 6-16 Ion Fragmentation Options Dialog Box  
8. Select the ion types to display, and whether or not to  
generate results for internal fragments.  
9. If you are analyzing Mariner data, select Calculate  
multiple charge states if desired, then specify the  
maximum charge state to calculate (up to 12).  
Masses that fall outside the tolerance are not labeled.  
11. To add the labels generated by this operation to the list  
of available user labels, select Append new labels to  
existing user label list. For more information, see  
Section 3.5.3, Setting Custom Peak Labels.  
12. Click OK.  
Data ExplorerSoftware Users Guide  
6-27  
 
Chapter 6 Using Tools and Applications  
Setting 13. Click User-Defined Amino Acids.  
user-defined amino  
acids  
6
The User-Defined Amino Acids dialog box (Figure 6-17)  
is displayed.  
Figure 6-17 User-Defined Amino Acids Dialog Box  
14. Add amino acid definitions and codes as needed.  
NOTE: You cannot modify pre-defined amino acids.  
User-defined amino acids are not saved when you  
close the Data Explorer software.  
15. Click Close.  
Calculating the 16. Click Induce Fragmentation.  
fragment ions  
Results are displayed in the Ion Fragmentation  
Calculator dialog box (Figure 6-18).  
6-28  
Applied Biosystems  
 
Using the Ion Fragmentation Calculator  
6
Figure 6-18 Ion Fragmentation Results for  
Synthetic Peptide (PPPPPPPPPPPPAR)  
Results Results are displayed in the:  
Ions tableLists the masses for each fragment and ion  
type.  
Internal fragments tableLists possible internal  
sequences, if you enabled this option in the Options  
dialog box (see Figure 6-16 on page 6-27).  
Click Clear Table Info to clear results. You can change  
options and recalculate ion fragmentation results.  
Data ExplorerSoftware Users Guide  
6-29  
   
Chapter 6 Using Tools and Applications  
Labeling peaks Click Label Peaks. The ion peaks specified in the Options  
dialog box are labeled on the trace if they are present  
(Figure 6-19).  
6
Hint: To screen out labels, decrease the Label Tolerance in  
the Options dialog box.  
Figure 6-19 Labeled Ion Fragmentation Peaks for  
Synthetic Peptide (PPPPPPPPPPPPAR)  
Hint: The Label Peaks function creates User Labels in the  
data file. To view, select Peak Label from the Peaks menu,  
then select User Label Setup. Click Save As to save the  
labels in a .LBS file for use with other data files. For more  
information, see Section 3.5.3, Setting Custom Peak  
Labels.  
To display the original labels, select Peak Label from the  
Peaks menu, then deselect User Labels.  
Creating a To create a PSD calibration reference file, click Create  
Reference File. For more information, see Creating PSD  
Calibration Reference (.REF) Fileson page 8-21.  
calibration  
reference file  
(.REF)  
6-30  
Applied Biosystems  
   
Using the Elemental Targeting Application  
6.6 Using the Elemental  
6
Targeting Application  
Description The Elemental Targeting application determines if observed  
masses in a spectrum correspond to chemical formulas you  
enter. This application generates a theoretical isotope pattern  
for the mass you enter using the Mass Resolution specified in  
Basic Peak Detection settings. It then compares each  
observed mass and isotope pattern to the theoretical mass  
and isotope pattern for each composition you entered, and  
reports an isotope match score that reflects how closely they  
match.  
Use the Elemental Targeting application to screen a spectrum  
for the presence of specific chemical compounds.  
Using the To use the Elemental Targeting application:  
Elemental  
Targeting  
application  
1. Display the spectrum of interest.  
2. Click the Spectrum window to activate it.  
3. From the Applications menu, select Elemental Targeting.  
The Elemental Targeting dialog box (Figure 6-20) is  
displayed.  
5. If the displayed Mass Peak Resolution is not appropriate  
for this calculation, change the setting in the Basic  
Settings tab of the Spectrum Peak Detection Setup dialog  
box. For more information, see Section 3.2.4, Peak  
Detection Parameter Descriptions.  
6. Enter formulas in the Element list by doing either of the  
following:  
Click  
, type a formula, select a charge state,  
then click OK.  
Click Import, then select a tab-delimited .TXT file  
that contains formulas and charge states.  
Data ExplorerSoftware Users Guide  
6-31  
         
Chapter 6 Using Tools and Applications  
Note the following when entering formulas:  
6
Spaces do not matter for formula. The first  
letter of two-letter elemental symbols must be  
capitalized (for example, Na).  
To ensure a better match between theoretical  
and observed isotopes, include the  
appropriate number of protons in the formula  
you enter for multiply charged ions.  
Figure 6-20 Elemental Targeting Dialog Box  
7. Click Calculate.  
6-32  
Applied Biosystems  
 
Using the Elemental Targeting Application  
Displaying results The results of the calculation are displayed in the Elemental  
Target tab of the Output window (Figure 6-21).  
6
Figure 6-21 Elemental Targeting Results in the Output Window  
Results include:  
IndexSequential number assigned to each result.  
FormulaElemental composition you entered.  
m/zMass/charge of an observed peak that, compared  
to the theoretical mass of the formula specified, is within  
the Mass Tolerance and Resolution you specified.  
ChargeCharge of the observed peak.  
Isotope Match ScoreNumber between 0.0 and 1.0  
that reflects how well the observed peak matches the  
theoretical formula, based on mass and isotope pattern.  
A higher number represents a better match.  
Isotope Match IntensityPeak area (counts) that  
overlaps between the observed isotope pattern and the  
theoretical isotope pattern. A higher number represents a  
better match.  
Displaying the To display the theoretical isotope distribution for a formula,  
double-click the line in the Elemental Target tab of the Output  
window.  
theoretical  
isotope  
The theoretical isotope trace is displayed in the Spectrum  
window with an ISO trace label and the elemental formula.  
distribution  
Data ExplorerSoftware Users Guide  
6-33  
             
Chapter 6 Using Tools and Applications  
6.7 Using the Macro Recorder  
6
Description The Macro Recorder feature in Data Explorer allows you to  
set up multi-step tasks to execute automatically when you  
click a macro button. The Macro Recorder feature creates a  
Visual Basic script of your actions as you record the macro,  
then executes the script when you run the macro.  
Macros provided Macros provided with the Data Explorer software are  
described in Appendix C, Data Explorer Toolbox (Visual Basic  
Macros).  
In this section This section includes:  
Before using the macro recorder  
Recording a macro  
Assigning macros to buttons  
Running a macro  
Deleting a macro  
Advanced macro editing  
Importing or Exporting Macros  
in DATAEXPLORER.VB6  
Running macros automatically when opening and  
closing files  
Importing macros Macros are provided with the Data Explorer software, but are  
not available for use until you import them into the Data  
Explorer project. For information, see Section 6.7.7, Importing  
or Exporting Macros in DATAEXPLORER.VB6.  
provided  
Maximum number You can create an unlimited number of macros. However,  
there are only 10 buttons in the toolbar to which you can  
assign macros at a given time. You can change the macro  
assigned to a button.  
of macros  
6-34  
Applied Biosystems  
         
Using the Macro Recorder  
Location of All macros you record are stored in a file called  
DATAEXPLORER.VB6 in the C:\MARINER\PROGRAM or  
C:\VOYAGER directory.  
macros  
6
Displaying the If the macro toolbar (Figure 6-22) is not displayed:  
macro toolbar  
1. Select Toolbar from the View menu.  
2. Select Macros, then click Close.  
Figure 6-22 Macro Toolbar  
If a numbered macro button is disabled (gray), no macro has  
been assigned to it. If a numbered macro button is enabled  
(green), a macro has been assigned to it.  
For a description of a toolbar button, place the cursor on the  
toolbar button. The name you assign to the macro or the  
default macro name is displayed below the button.  
Functions not Most commands in the Data Explorer software can be  
included in a macro that you create with the Macro Recorder.  
supported  
The following menu commands are not supported by the  
Macro Recorder:  
Menu  
Commands Not Supported  
File  
Open  
Close  
Close All  
Result SpectrumDelete  
Convert All SpectraNew Data Format  
ExportResult Spectrum  
SettingsSave Processing Settings  
SettingsRestore Graphic Settings  
SettingsSave Graphic Settings  
SettingsRevert to last saved Graphic/Processing Settings  
Print Setup  
Exit  
Continued  
Data ExplorerSoftware Users Guide  
6-35  
           
Chapter 6 Using Tools and Applications  
Menu  
View  
Display  
Commands Not Supported  
No commands supported  
Add/Remove Traces  
6
NOTE: The  
and  
buttons are supported by the Macro Editor.  
Processing History  
Range  
Graphic Options*  
DefaultRevert to Last Graphic Settings  
Process  
PSD processing  
Mass CalibrationManual Calibration  
Mass CalibrationPSD calibration  
Mass CalibrationAutomatic Calibration  
Mass CalibrationRevert to instrument calibration  
Mass CalibrationEdit/Create Reference File  
Dual Spectral Trace Arithmetic  
Multiple ChargeMass Deconvolution  
Peaks  
Tools  
Peak Detection*  
Peak Label*  
Insert Peak (chromatogram)  
Processing History Options  
Customize Toolbar  
Customize ToolMenu  
Macro commands, including Automatic Macros  
Applications  
Window  
Elemental Targeting  
Ion Fragmentation Calculator  
No commands supported  
* You can apply specific Graphic Options or Peak Detection settings by creating a macro that loads a  
.SET file containing the desired settings. You can apply specific peak labels by creating a macro that  
loads a .USR (user label file) containing the desired labels.  
Functions you perform in the Output window, for example,  
sorting or copying the peak list, are not supported by the  
Macro Recorder.  
6-36  
Applied Biosystems  
Using the Macro Recorder  
6.7.2 Recording a Macro  
6
To record a macro:  
1. Open a data file.  
2. From the Tools menu, select Record New Macro.  
The Record Macro dialog box (Figure 6-23) is  
displayed.  
Figure 6-23 Record Macro Dialog Box  
3. Type a name and a description if desired.  
4. Click OK.  
macro. For example, select Noise Filter/Smooth from the  
Process menu, select the smoothing method, specify the  
associated parameter, then click OK.  
Not all functions in Data Explorer are supported. See  
Functions not supportedon page 6-35.  
6. From the Tools menu, select Stop Macro Recording.  
The macro is saved and added to the  
DATAEXPLORER.VB6 file.  
Data ExplorerSoftware Users Guide  
6-37  
     
Chapter 6 Using Tools and Applications  
6.7.3 Assigning Macros to Buttons  
6
Only macros present in the DATAEXPLORER.VB6 file can be  
assigned to buttons and run in the Data Explorer software.  
NOTE: If you have installed a new version of Data Explorer  
software, new macros may be provided. New macros are  
not available for use until you import them into the Data  
Explorer project. For information, see Section 6.7.7,  
Importing or Exporting Macros in DATAEXPLORER.VB6.  
Assigning a You can assign macros to 10 buttons. To assign a macro:  
macro to a button  
1. From the Tools menu, select Assign Macro.  
The Assign Macro dialog box (Figure 6-24) is  
displayed.  
Figure 6-24 Assign Macro Dialog Box  
2. Select the button to assign from the list on the left, and the  
macro to assign from the list on the right.  
3. Click Assign.  
The macro button in the toolbar to which you assigned  
the macro turns from gray to green (if it did not  
previously have a macro assigned to it).  
6-38  
Applied Biosystems  
       
Using the Macro Recorder  
De-assigning a To de-assign a macro from a button, select the macro button in  
the Assign Macro dialog box, then click De-assign.  
macro from a  
6
button  
6.7.4 Running a Macro  
You can run a macro using a:  
Toolbar button  
Menu command  
Using a toolbar To run a macro using the toolbar button:  
button  
1. Open the data file on which you want to run the macro.  
2. Click the button assigned to the macro you want to run.  
Hint: Place the cursor on a macro button to display the  
name of the macro assigned to the button.  
The macro executes.  
Using the menu To run a macro using the menu:  
command  
1. Open the data file on which you want to run the macro.  
2. From the Tools menu, select Macros.  
The Macros dialog box (Figure 6-25) is displayed.  
Data ExplorerSoftware Users Guide  
6-39  
   
Chapter 6 Using Tools and Applications  
6
Figure 6-25 Macros Dialog Box  
3. Select the macro to run from the list.  
4. Click Run.  
If the macro If the macro contains a syntax error, it may cause the Data  
Explorer software to close unexpectedly. If this occurs, restart  
contains a syntax  
the Data Explorer software and examine the macro in the  
Visual Basic Editor. See Section 6.7.6, Advanced Macro  
Editing.  
error  
Macro lines containing syntax errors are displayed in color in  
the Visual Basic Editor. For information on correcting errors,  
refer to the online help available in the Visual Basic Editor.  
6-40  
Applied Biosystems  
 
Using the Macro Recorder  
6.7.5 Deleting a Macro  
6
To delete a macro:  
1. From the Tools menu, select Macros.  
The Macros dialog box (Figure 6-26) is displayed.  
Figure 6-26 Macros Dialog Box  
2. Select the macro to delete from the list.  
3. Click Delete.  
NOTE: Other buttons on this dialog box are for  
advanced editing. Refer to the online help available  
within the Visual Basic Editor.  
Data ExplorerSoftware Users Guide  
6-41  
     
Chapter 6 Using Tools and Applications  
6.7.6 Advanced Macro Editing  
6
Accessing the You can access the Visual Basic Editor to enhance or edit a  
script created by the Macro Recorder in Data Explorer, or to  
create a new script.  
Visual Basic  
Editor  
Access the Visual Basic Editor from Data Explorer in three  
ways:  
Select Visual Basic Editor from the Tools menu.  
Select Macros from the Tools menu, select the macro to  
edit, then click Edit.  
Click  
in the Macro toolbar  
The DATAEXPLORER.VB6 file is displayed in the Visual Basic  
Project Explorer window.  
Macro scripts created using the Macro Recorder in Data  
Explorer are stored in Module folders in the Project Explorer  
window.  
Displaying To display a list of the scriptable objects available within Data  
Explorer, select Object Browser from the View menu in the  
Visual Basic Project Explorer window.  
scriptable objects  
For information on using the Visual Basic Editor, refer to the  
online help in the Visual Basic Editor.  
Numbering When you create Macro scripts, the Visual Basic Editor uses a  
numbering sequence that starts at 0. If you specify a number  
sequence  
of 1 in the Visual Basic Editor, it indicates the second item in a  
sequence, not the first. For example, if you create a script to  
sort the Peak List in the Output window, and you want to sort  
by Area (column 8 in the Peak List), specify sorting on column  
7 in the script.  
6-42  
Applied Biosystems  
   
Using the Macro Recorder  
6.7.7 Importing or Exporting Macros  
in DATAEXPLORER.VB6  
6
You can import macros into, or export macros from, the  
DATAEXPLORER.VB6 project for use in the Data Explorer  
software.  
Importing when When you install a new version of the Data Explorer software:  
new versions of  
Data Explorer  
software installed  
The DATAEXPLORER.VB6 file is not overwritten. This  
allows you to maintain any macros you have developed  
in the DATAEXPLORER.VB6 file.  
New macros are supplied in stand-alone ASCII format  
(.BAS and .FRM files) in C:\MARINER\PROGRAM\  
MACROS or C:\VOYAGER\MACROS and are not  
incorporated into the DATAEXPLORER.VB6 file  
automatically.  
To make the new macros available for use, import them into  
the DATAEXPLORER.VB6 file.  
Importing NOTE: You can also import by opening NT Explorer, then  
displayed in the Visual Basic Editor window.  
To import macros:  
1. Access the Visual Basic Editor from Data Explorer as  
described in Section 6.7.6, Advanced Macro Editing.  
2. From the File menu, select Import File.  
3. Select the .BAS file (Basic File module containing the  
macros) or .FRM file (form containing the user interface  
for the macros, if applicable), then click Open.  
Data ExplorerSoftware Users Guide  
6-43  
         
Chapter 6 Using Tools and Applications  
The selected macros are imported into the  
DATAEXPLORER.VB6 project. The .BAS files are  
6
DataExplorerProject, and the .FRM files are included in  
the Forms folder in the DataExplorerProject. All macros  
imported into the DataExplorerProject are displayed in  
the list of macros you can assign in the Data Explorer  
software. See Section 6.7.3, Assigning Macros to  
Buttons.  
Exporting To export macros:  
1. Access the VBA Editor from Data Explorer as described  
above.  
2. From the File menu, select Export File.  
3. Select the .BAS or .FRM file to export, then click Save.  
NOTE: You have to export modules or forms one at  
a time.  
6-44  
Applied Biosystems  
 
Using the Macro Recorder  
6.7.8 Running Macros Automatically  
You can set the Data Explorer software to automatically run  
macros you previously created when you open or close a data  
file. For information on creating macros, see Section 6.7.2,  
Recording a Macro.  
To set up automatic macros:  
1. From the Tools menu, select Automatic Macros.  
The Automatic Macro Setup dialog box (Figure 6-26) is  
displayed.  
Figure 6-27 Automatic Macro Setup Dialog Box  
2. Select File Open Macro or File Close Macro, as desired,  
then select the macro to run when you open or close a  
data file.  
3. Click OK.  
Data ExplorerSoftware Users Guide  
6-45  
         
Chapter 6 Using Tools and Applications  
6
6-46  
Applied Biosystems  
Chapter  
7
7 Data Explorer  
Examples  
This chapter contains the following sections:  
7.1  
Mariner Data Examples ...................................... 7-2  
7.1.1  
7.1.2  
Improving Signal-To-Noise Ratio ......... 7-2  
Deconvoluting and Evaluating  
Unresolved Chromatographic Peaks ... 7-4  
7.1.3  
Determining if a Peak is  
Background Noise .............................. 7-8  
7.2  
Voyager Data Examples..................................... 7-11  
7.2.1  
7.2.2  
7.2.3  
Detecting and Labeling  
Partially Resolved Peaks ...................7-11  
Processing Before Calibrating to  
Optimize Mass Accuracy .................. 7-14  
Detecting Peaks from  
Complex Digests ............................... 7-18  
Data ExplorerSoftware Users Guide  
7-1  
     
Chapter 7 Data Explorer Examples  
7.1 Mariner Data Examples  
This section includes:  
Improving signal-to-noise ratio  
Deconvoluting and evaluating unresolved  
chromatographic peaks  
Determining if a peak is background noise  
7
7.1.1 Improving Signal-To-Noise Ratio  
Overview You can improve the signal-to-noise ratio for low-level  
components in the total ion chromatogram (TIC) by creating  
an extracted ion chromatogram for the mass of interest.  
Each data point in the TIC represents the sum of all ion  
intensities for all masses in the corresponding spectrum. By  
creating an extracted ion chromatogram for a specific mass,  
you can screen out ion contribution from masses that you are  
not interested in.  
Creating an The following example illustrates how you can create an  
extracted ion chromatogram to improve the signal-to-noise  
extracted ion  
chromatogram  
1. Display the data file.  
2. In the Chromatogram window, click  
in the toolbar, or  
set the Trace Replace mode to Add, if you want to add a  
new trace. See Section 2.4.4, Adding Traces from the  
Same Data File to a Window, for information.  
3. From the Display menu, select Trace, then select  
Extracted Ion.  
4. In the Extracted Ion Chromatogram window, select  
Center/Window, then type the mass of interest (609 Da)  
and the window for the mass (for example, 0.5 Da).  
5. Click OK.  
7-2  
Applied Biosystems  
             
Mariner Data Examples  
Figure 7-1 illustrates the improved signal-to-noise ratio  
in the extracted ion chromatogram for three replicate  
loop injections.  
7
Original TIC  
containing all  
masses  
Extracted ion  
chromatogram for 609 Da  
with improved  
signal-to-noise ratio  
Figure 7-1 Improving Signal-To-Noise Ratio with an  
Extracted Ion Chromatogram  
Data ExplorerSoftware Users Guide  
7-3  
   
Chapter 7 Data Explorer Examples  
7.1.2 Deconvoluting and  
Evaluating Unresolved  
Chromatographic Peaks  
Overview You can use the Data Explorer software to deconvolute  
chromatographic peaks and obtain masses for each  
component by:  
Creating a combined spectrum to identify the coeluting  
species by mass  
7
Creating extracted ion chromatograms for the masses,  
and comparing the extracted ion chromatograms to the  
original TIC containing the unresolved peaks  
You can then create a combined spectrum from each  
extracted ion chromatogram to evaluate the spectral data.  
Creating a The following example illustrates how to deconvolute two  
unresolved chromatographic peaks in a tryptic digest of  
cytochrome c:  
combined  
spectrum  
1. Display the data file.  
2. In the Chromatogram window, zoom in on the unresolved  
peaks (Figure 7-2).  
Figure 7-2 Unresolved Peaks in Cytochrome C  
3. In the Chromatogram window, right-click-drag over the  
unresolved peak pair.  
7-4  
Applied Biosystems  
           
Mariner Data Examples  
The combined spectrum is displayed (Figure 7-3), with  
two intense peaks at 410 Da and 723 Da. Generate  
extracted ion chromatograms as described below to  
determine if these peaks are the coeluting components.  
NOTE: If these peaks do not yield chromatograms with  
profiles that correspond to the unresolved peaks, try  
another spectral peak.  
7
Figure 7-3 Combined Spectrum for Unresolved Peaks in Cytochrome C  
Creating To create extracted ion chromatograms:  
extracted ion  
1. With the Chromatogram window activated, click  
in  
chromatograms  
the toolbar two times to add two traces. Click an added  
trace to activate it.  
2. In the Spectrum window, right-click-drag over the first  
peak.  
The extracted ion chromatogram is displayed.  
3. In the Chromatogram window, click the second added  
trace to activate it.  
4. In the Spectrum window, right-click-drag over the second  
peak.  
The extracted ion chromatogram is displayed.  
Figure 7-4 illustrates the peaks deconvoluted from the  
original unresolved peaks.  
Data ExplorerSoftware Users Guide  
7-5  
   
Chapter 7 Data Explorer Examples  
7
Original TIC  
containing  
unresolved peaks  
Extracted ion  
chromatograms with  
deconvoluted peaks  
Figure 7-4 Deconvoluting Unresolved Chromatographic Peaks  
Creating Create a combined spectrum for each extracted ion  
chromatogram:  
combined spectra  
1. Activate the Spectrum window, then click  
toolbar two times to add two traces.  
in the  
2. Right-click-drag over the first half of the mass range in the  
410 extracted ion chromatogram.  
3. Activate the second trace in the Spectrum window.  
4. Right-click-drag over the second half of the mass range in  
the 723 extracted ion chromatogram.  
7-6  
Applied Biosystems  
   
Mariner Data Examples  
Figure 7-5 illustrates the combined spectra for the  
deconvoluted peaks.  
Note that both spectra contain a peak at 391 Da which  
requires investigation to determine if it is a low-level  
component or background noise. See Section 7.1.3,  
Determining if a Peak is Background Noise.  
7
Both combined spectra  
contain a peak at 391 Da  
Figure 7-5 Combined Spectra for Deconvoluted Peaks  
Data ExplorerSoftware Users Guide  
7-7  
 
Chapter 7 Data Explorer Examples  
7.1.3 Determining if a Peak  
is Background Noise  
Overview To determine if spectral peaks represent low-level  
components or if they are due to solvent contribution, you can:  
Create an extracted ion chromatogram for the spectral  
peaks  
7
Subtracting The following example uses the chromatograms and spectra  
from Section 7.1.2, Deconvoluting and Evaluating Unresolved  
spectral peaks  
Chromatographic Peaks. It illustrates how to subtract baseline  
and determine if the spectral peak observed at 391 Da  
(Figure 7-5 on page 7-7) is eliminated.  
To subtract spectra:  
1. Activate the first extracted ion chromatogram in  
Chromatogram window (see Figure 7-4 on page 7-6).  
2. From the Process menu, select Add/Subtract Spectra.  
3. Click  
, then right-click-drag over the peak in the  
extracted ion chromatogram.  
The spectrum range is displayed in the Spectra To Be  
Added list in the Add and Subtract Spectra dialog box  
(Figure 7-6).  
4. In the Add and Subtract Spectra dialog box, click  
anywhere in the Spectra To Be Subtracted list.  
5. In the TIC, right-click over a region of baseline that does  
not contain any peaks.  
NOTE: When selecting spectra to subtract, select from  
the TIC. The baseline in the extracted ion  
chromatograms is for a selected mass and does not  
represent all components which may be present.  
7-8  
Applied Biosystems  
       
Mariner Data Examples  
The spectrum range is displayed in the Spectra To Be  
Subtracted list in the Add and Subtract Spectra dialog  
box (Figure 7-6).  
7
Figure 7-6 Subtracting Spectra  
6. Click OK.  
The subtracted spectrum is displayed (Figure 7-7 on  
page 7-10).  
Data ExplorerSoftware Users Guide  
7-9  
 
Chapter 7 Data Explorer Examples  
Figure 7-7 Subtracted Spectrum  
7
The peak at 391 is still present, which indicates one of the  
following conditions:  
You did not subtract sufficient baseline  
The peak is a coeluting component  
To evaluate further, create an extracted ion chromatogram for  
391 Da as described below and evaluate the signal.  
Creating To create the extracted ion chromatogram, right-click-drag  
over the peak at 391 Da in the spectrum trace.  
extracted ion  
chromatogram  
NOTE: You can also create an extracted ion chromatogram  
by selecting Trace from the Display menu, selecting  
Extracted Ion, then entering a mass.  
Figure 7-8 illustrates the extracted ion chromatogram for mass  
391. The signal is relatively consistent, indicating that it is  
background instead of a component peak.  
Figure 7-8 Extracted Ion Chromatogram for Mass 391 Da  
7-10  
Applied Biosystems  
     
Voyager Data Examples  
7.2 Voyager Data Examples  
This section includes:  
Detecting and labeling partially resolved peaks  
Processing before calibrating to optimize mass accuracy  
Detecting peaks from complex digests  
7.2.1 Detecting and Labeling  
7
Partially Resolved Peaks  
If peaks are not If peaks are partially resolved and the peaks of interest are not  
labeled, you can adjust the following peak detection  
parameters:  
labeled  
If...  
Adjust the following...  
Peaks represent two  
compounds, and you  
want both peaks labeled  
In the Peak Label dialog box, select Allow Overlapping  
Peak Labels.  
OR  
On the Basic Settings tab in Peak Detection, set %Max  
Peak Area to 0, then adjust the %Base Peak Intensity  
until peaks are detected.  
OR  
On the Peak Processing tab in Peak Detection, change  
the default Integration Baseline Setting from  
Valley-to-Valley to Valley-to-Baseline.  
Peaks represent partially On the Basic Settings tab in Peak Detection, decrease  
resolved isotopes, and  
you want to label and  
detect the average mass  
the Mass Resolution setting until the isotopic envelope is  
detected.  
Figure 7-9 shows a spectrum with partially resolved peaks that  
represent two compounds.  
Data ExplorerSoftware Users Guide  
7-11  
             
Chapter 7 Data Explorer Examples  
Figure 7-9 Partially Resolved Peaks That Represent Two Compounds, Minor  
Component Not Detected  
7
Adjusting peak To adjust peak detection:  
detection  
1. Click  
in the toolbar or select Peak Detection from  
the Peaks menu.  
The Spectrum Peak Detection Setup dialog box is  
displayed with the Basic Settings tab (Figure 7-10)  
displayed.  
Figure 7-10 Spectrum Peak Detection Setup—  
Basic Settings Tab  
7-12  
Applied Biosystems  
   
Voyager Data Examples  
2. Do either of the following:  
Change the %Max Peak Area from 1 (the default)  
to 0, and the %Base Peak Intensity from 0 (the  
default) to 1.  
Click the Peak Processing tab and change the  
default Integration Baseline Setting from  
Valley-to-Valley to Valley-to-Baseline.  
3. Click OK.  
7
Figure 7-11 shows the partially resolved peaks that are now  
labeled.  
Figure 7-11 Partially Resolved Peaks Labeled  
Data ExplorerSoftware Users Guide  
7-13  
 
Chapter 7 Data Explorer Examples  
7.2.2 Processing Before Calibrating to  
Optimize Mass Accuracy  
This section includes:  
Calibrating without baseline correcting and deisotoping  
Before calibrating  
Calibrating  
Calibrating For optimum mass accuracy, baseline correct and deisotope a  
without baseline  
7
spectrum before calibrating.  
correcting and  
deisotoping  
Figure 7-12 shows a spectrum before calibration.  
Figure 7-12 Spectrum Before Calibration  
Figure 7-13 shows a spectrum that has been calibrated  
without initial baseline correction and deisotoping. Mass  
accuracy is not acceptable.  
Figure 7-13 Calibration Without Baseline Correction  
or Deisotoping  
7-14  
Applied Biosystems  
       
Voyager Data Examples  
Before calibrating To optimize mass accuracy, do the following before calibrating:  
1. Display the spectrum of interest.  
Baseline correcting 2. From the Process menu, select Baseline Correction.  
The spectrum is baseline corrected. For more  
information, see Section 5.8.2, Using Baseline  
Correction.  
Deisotoping 3. From the Peaks menu, select Peak Deisotoping.  
7
The Deisotoping dialog box (Figure 7-14) is displayed.  
Figure 7-14 Deisotoping Dialog Box  
4. For this example spectrum, specify H for Adduct and  
C6H5NO for Generic Formula.  
5. Click OK.  
Figure 7-15 shows the baseline-corrected, deisotoped  
spectrum before calibration. For more information on  
deisotoping, see Section 3.4, Deisotoping a Spectrum.  
Figure 7-15 Baseline-Corrected, Deisotoped Spectrum  
Before Calibration  
Data ExplorerSoftware Users Guide  
7-15  
   
Chapter 7 Data Explorer Examples  
Calibrating To calibrate the deisotoped spectrum:  
1. From the Peaks menu, select Peak Label, and select  
the Mass Label Type (peak apex or peak centroid) to  
use for calibration. Click OK.  
2. From the Process menu, select Mass Calibration and  
then select Manual Calibration.  
The Manual Calibration dialog box is displayed  
(Figure 7-16).  
7
Figure 7-16 Manual Calibration Dialog Box  
3. Click  
, then select the VOYAGER.REF calibration  
4. Enter Reference Matching and Calibration Criteria.  
NOTE: For descriptions of calibration parameters, see  
Section 5.3, Manual Calibration.  
7-16  
Applied Biosystems  
 
Voyager Data Examples  
Matching peaks 5. Click Match Peaks and Solve. The software compares  
observed masses in the spectrum to reference masses in  
the selected reference file, lists the matches in the Peak  
Matched list, calibrates the spectrum, and displays the  
calibration statistics in the Output window.  
NOTE: If you set Mass Tolerance too low, no peaks will  
match.  
7
You can also add peaks to the Peak Matched list by  
right-click-dragging on a peak in the spectrum, then  
selecting the mass from the reference Peak Information  
dialog displayed. Click Solve and Plot after manually  
adding masses.  
The spectrum is calibrated and displayed with an MC  
trace label. The calibration statistics are displayed in  
the Output window.  
Applying 6. If the calibration statistics are acceptable, click Apply  
Calibration.  
Figure 7-17 shows the deisotoped spectrum after calibration,  
with improved mass accuracy.  
Figure 7-17 Deisotoped Spectrum After Calibration  
Data ExplorerSoftware Users Guide  
7-17  
     
Chapter 7 Data Explorer Examples  
7.2.3 Detecting Peaks  
from Complex Digests  
Overview Complex digests often contain hundreds of peaks which may  
have relatively low signal-to-noise ratios. To quickly screen out  
noise and detect peaks of interest:  
Noise filter/smooth to remove initial noise.  
Set initial peak detection thresholds low enough to detect  
all peaks, to ensure that monoisotopic peaks are detected  
for proper deisotoping.  
7
Deisotope to identify isotope peak clusters. The  
deisotope function amplifies peak intensity based on the  
total area of all peak areas in the isotope cluster. It does  
not amplify noise peaks.  
Increase peak detection thresholds to eliminate noise  
peaks.  
Figure 7-18 represents a spectrum trace from a digest.  
Overlapping Peak Labels are enabled to illustrate the large  
number of peaks detected in the spectrum (over 300 peaks  
contained in the peak list displayed in the Output window).  
Figure 7-18 Digest with Several Hundred Peaks  
Detected (Allow Overlapping Labels Enabled)  
7-18  
Applied Biosystems  
       
Voyager Data Examples  
Procedure To detect peaks from complex mixtures:  
1. Display the spectrum of interest.  
Noise 2. From the Process menu, select Noise Filter/Smooth.  
filtering/smoothing  
The Noise Filter/Smooth dialog box (Figure 7-19) is  
displayed.  
7
Figure 7-19 Noise Filter/Smooth Dialog Box  
3. Select Default Smoothing or Noise Filter (with a  
Correlation Factor of 0.7), then click OK.  
For more information, see Section 5.7, Noise  
Filtering/Smoothing.  
4. Click  
in the toolbar or select Peak Detection from  
Setting detection  
thresholds  
the Peaks menu.  
The Spectrum Peak Detection Setup dialog box is  
displayed with the Basic Settings tab (Figure 7-20)  
displayed.  
Data ExplorerSoftware Users Guide  
7-19  
 
Chapter 7 Data Explorer Examples  
7
Figure 7-20 Spectrum Peak Detection Setup—  
Basic Settings Tab  
5. Click Use Advanced Settings.  
The Advanced Settings tab is displayed (Figure 7-21).  
7-20  
Applied Biosystems  
 
Voyager Data Examples  
7
Figure 7-21 Spectrum Peak Detection Setup—  
Advanced Settings Tab  
6. Set Minimum Area to 0.  
7. Click the Basic Settings tab (see Figure 7-20 on  
page 7-20), then set %Max Peak Area to 0 or 0.1.  
8. Click Apply. Note that many peaks are added to the peak  
list in the Output window.  
For more information, see Section 3.2, Peak Detection.  
Deisotoping 9. From the Peaks menu, select Peak Deisotoping.  
The Deisotoping dialog box (Figure 7-14) is displayed.  
Data ExplorerSoftware Users Guide  
7-21  
 
Chapter 7 Data Explorer Examples  
7
Figure 7-22 Deisotoping Dialog Box  
10. For this example spectrum, specify H for Adduct and  
11. Click OK. For more information on deisotoping, see  
Section 3.4, Deisotoping a Spectrum.  
Increasing 12. In the Basic Settings tab in Global Thresholds (see  
Figure 7-20 on page 7-20), increase %Max Peak Area  
to a setting between 0.1 and 1% until the peaks of  
interest are detected and noise is screened out.  
detection  
thresholds  
Figure 7-23 shows the trace after deisotoping. Note  
that many fewer peaks are detected. By further  
fine-tuning %Max Peak Area, you can screen out  
additional noise peaks.  
Figure 7-23 Digest After Noise Filtering, Deisotoping,  
and Adjusting Detection Thresholds  
7-22  
Applied Biosystems  
 
Chapter  
8
8 Viewing  
Voyager PSD Data  
This chapter contains the following sections:  
8.1  
8.2  
8.3  
Displaying PSD Data ....................................... 8-2  
Applying Fragment Labels ............................... 8-8  
Calibrating a PSD Spectrum .......................... 8-10  
8.3.1  
8.3.2  
8.3.3  
Checking Peak Detection ............... 8-11  
Calibrating .....................................8-12  
Creating PSD .CAL Files and  
Applying to Other Data Files ..........8-20  
8.3.4  
8.3.5  
Creating PSD Calibration  
Reference (.REF) Files ..................8-21  
Changing the Precursor Mass ........8-23  
Data ExplorerSoftware Users Guide  
8-1  
   
Chapter 8 Viewing Voyager PSD Data  
8.1 Displaying PSD Data  
This section includes:  
Displaying the composite spectrum  
Advancing through segment traces  
Displaying multiple segment traces  
Redisplaying the composite spectrum  
How the composite spectrum is generated  
NOTE: For information on acquiring PSD spectra, see the  
Voyager Biospectrometry Workstation Users Guide.  
Displaying the NOTE: You cannot display the PSD data you are currently  
acquiring in the Instrument Control Panel in the Data  
Explorer software until you stop the PSD experiment in the  
Instrument Control Panel. Error messages are displayed if  
you try to open the file. For more information, see the  
Voyager Biospectrometry Workstation Users Guide.  
composite  
spectrum  
8
To display PSD data:  
1. Open the PSD data file of interest in the Data Explorer  
software as described in Section 2.1, Opening and  
Closing Data Files.  
NOTE: PSD data files are named with a .DAT  
example, Test_PSD, to help you distinguish them from  
non-PSD data files.  
The composite spectrum is generated as described in  
How the composite spectrum is generatedon  
page 8-6, and displayed with a Stitched PSD trace label  
(Figure 8-1).  
8-2  
Applied Biosystems  
                 
Displaying PSD Data  
8
Figure 8-1 PSD Spectrum in Data Explorer  
Advancing To advance through segment traces, click  
through segment  
traces  
Segments are displayed in the order in which they were  
acquired.  
NOTE: If these buttons are not displayed in the toolbar, you  
can add them. See Section 1.4.3, Customizing Toolbars.  
Displaying To display multiple segment traces:  
multiple segment  
traces  
1. Click  
in the toolbar three times to add three Not  
Usedtraces.  
2. From the Process menu, select PSD Processing.  
Data ExplorerSoftware Users Guide  
8-3  
   
Chapter 8 Viewing Voyager PSD Data  
The PSD Processing dialog box is displayed  
(Figure 8-2) and lists all segments contained in the PSD  
.DAT file in the order in which they were acquired with  
associated Mirror Ratios and Max Stitch Masses.  
Column  
header  
8
Figure 8-2 PSD Processing Dialog Box  
The Max Stitch Mass is equal to the Precursor Mass  
times the Mirror Ratio. This value reflects the maximum  
mass of the segment that will be included in the composite  
spectrum. The mass range included in the segment is  
approximately 15 percent higher than the Max Stitch  
Mass. Optimum focus and resolution is achieved for  
fragment ions close to this mass.  
8-4  
Applied Biosystems  
         
Displaying PSD Data  
NOTE: The entry number in the PSD Segment list  
above may not correspond to Segment number  
specified in the Segment list for acquisition (described  
in the Voyager Biospectrometry Workstation Users  
Guide) if you did not acquire all segments, or did not  
acquire segments in order of decreasing Mirror Ratio.  
The number above reflects the order of acquisition. For  
example, if you listed Segments 1 through 5 for  
acquisition, but acquired Segments 1, 3, and 5, you  
would see three entries above that correspond to the  
three acquired segments.  
3. In the Spectrum window, click the first Not Used trace.  
4. In the PSD Processing dialog box, double-click the Entry  
number of the segment to add.  
5. Repeat step 3 and step 4 to display additional segments.  
6. To advance through traces, select any trace, then click  
and  
.
Redisplaying the To redisplay the composite spectrum, click Generate  
Composite in the PSD Processing dialog box (Figure 8-2).  
composite  
spectrum  
Data ExplorerSoftware Users Guide  
8-5  
     
Chapter 8 Viewing Voyager PSD Data  
How the The software does the following to generate a composite  
spectrum:  
composite  
spectrum is  
generated  
Evaluates all segments in the .DAT file to determine if  
there are multiple segments acquired using the same  
If there are duplicates, selects the most recently acquired  
segment to include in the composite spectrum.  
Using the precursor ion mass and calibration constants  
from the PSD calibration in the .DAT file (see PSD  
calibration equationbelow), determines the region of  
each segment to include in the composite spectrum as  
illustrated in Figure 8-3 on page 8-7.  
Applies baseline correction and two-standard-deviation  
noise removal to the composite region of each segment.  
• “Stitchesthe composite regions together and displays  
the composite spectrum.  
8
PSD calibration The equation that the Voyager software uses to calculate the  
mass for fragment ions is:  
equation  
α Ri + β  
α + β  
2
mp  
mf  
=
(α R + β)  
[(tf tp) + γ (tf tp) ] +  
i
where:  
m
=
=
=
=
Fragment ion mass  
f
Calibration constants  
Mirror Ratio  
α, β, γ  
R
i
t
t
Fragment ion flight time  
f
=
Precursor ion flight time (at R = 1) calculated  
p
i
during acquisition  
m
=
Precursor ion mass entered in PSD Acquisition settings  
before acquisition, or changed in PSD Processing  
p
8-6  
Applied Biosystems  
         
Displaying PSD Data  
If you are performing an internal standard calibration, the  
software determines the constants as listed below:  
PSD Internal  
Standard Calibration  
Constant  
Value Used  
One-point  
α
Calculated from standard mass  
β and γ  
α and β  
γ
0
Two-point or  
three-point  
Calculated from standard masses  
0
More than three-point α, β, γ  
Calculated from standard masses  
Region of The composite spectrum is generated from portions of the  
segment traces. The upper mass limit of the composite region  
8
segments included  
in each segment is determined by the PSD Mirror Ratio (Rn)  
with which the segment was acquired and the mass of the  
precursor ion (mp) (Figure 8-3).  
in composite  
spectrum  
included in composite  
included in composite  
Seg 1  
Seg 2  
R x m  
1
p
included in composite  
(If first segment acquired  
with PSD Mirror Ratio=1.0,  
upper limit is  
slightly higher than  
precursor ion mass)  
Seg 3  
Seg 4  
R x m  
2
p
R x m  
3
p
Figure 8-3 Portions of Segment Traces Included in the Composite Spectrum  
Data ExplorerSoftware Users Guide  
8-7  
   
Chapter 8 Viewing Voyager PSD Data  
8.2 Applying Fragment Labels  
Overview Use the Ion Fragmentation calculator to apply fragment labels.  
For detailed information on using the Ion Fragmentation  
calculator, see Section 6.5, Using the  
Ion Fragmentation Calculator.  
Applying labels To apply fragment labels to PSD spectra:  
1. From the Applications menu, select Ion Fragmentation  
Calculator.  
The Ion Fragmentation Calculator dialog box  
(Figure 8-4) is displayed.  
8
Figure 8-4 Ion Fragmentation Calculator Dialog Box  
8-8  
Applied Biosystems  
             
2. In the Sequence text box, type the amino acid sequence  
of the compound. Use single-letter codes. Set other  
parameters as needed. For parameter descriptions, see  
Section 6.5, Using the Ion Fragmentation Calculator.  
3. Click Options to specify the fragment peaks to label.  
4. Click Induce Fragmentation.  
5. Click Label Peaks.  
The ion peaks specified in Options are labeled on the  
trace if they are present.  
Hint: To more selectively apply labels, decrease the  
Mass Tolerance in the Options dialog box.  
Hint: This function creates User Labels in the data file.  
To view, select Peak Label from the Peaks menu, then  
click User Label Setup. Click Save As to save the  
labels in a .LBS file for use with other data files. For  
more information, see Section 3.5.3, Setting Custom  
Peak Labels.  
8
6. To display the original labels, select Peak Label from the  
Peaks menu, then deselect User Labels.  
Data ExplorerSoftware Users Guide  
8-9  
 
Chapter 8 Viewing Voyager PSD Data  
8.3 Calibrating a PSD Spectrum  
NOTE: Multi-point calibration yields higher mass accuracy  
than one-point calibration.  
This section includes:  
Checking peak detection  
Calibrating  
Creating PSD .CAL files and applying to other data files  
Creating PSD calibration reference (.REF) files  
Changing the precursor mass  
When to use this Use this procedure to:  
procedure  
Generate a PSD calibration (.CAL) file from a known  
standard analyzed in PSD mode. You can then export the  
PSD .CAL file and include the PSD .CAL file in a PSD  
instrument settings (.BIC) file you use to acquire unknown  
PSD samples.  
8
Internally calibrate an unknown PSD data file by  
specifying known monoamino acid fragment ions as the  
Overview of To create a PSD .CAL file:  
creating a PSD  
.CAL file  
Check peak detection as described in Section 8.3.1,  
Create a calibration reference (.REF) file as described in  
Calibration Reference (.REF) Files.  
Internally calibrate the PSD .DAT file as described in  
Section 8.3.2, Calibrating. Specify the calibration  
reference (.REF) file created in Section 8.3.4, Creating  
PSD Calibration Reference (.REF) Files, when  
calibrating.  
Export the calibration (.CAL) file from the PSD .DAT file to  
use when acquiring unknown samples in PSD mode.  
8-10  
Applied Biosystems  
             
Calibrating a PSD Spectrum  
8.3.1 Checking Peak Detection  
Checking Before calibrating, check that peaks in all segment traces of  
interest are properly peak detected and that noise is not  
detected as peaks. Note the following when setting peak  
detection parameters:  
Peak detection settings are applied to the currently  
displayed composite spectrum or segment traces.  
The default peak detection settings (in  
VOYAGERPSD.SET) have different detection ranges for  
different mass ranges. You can fine tune ranges and  
Filter Width settings in ranges to optimize detection. For  
more information, see Detection Rangeson page 3-28.  
Fragment ion peaks in segments collected with lower  
PSD Mirror Ratios are broader and include more data  
points. Use a higher Filter Width setting for these  
segments.  
8
Within a segment, resolution increases with increasing  
mass, as the flight time of the fragment ions approaches  
the flight time of the precursor. Use a lower Filter Width  
setting for higher resolutions.  
For more information, see Section 3.2, Peak Detection.  
Smoothing If a segment appears noisy, smooth the trace (for example,  
use 25-point Gaussian smoothing) before calibrating.  
Data ExplorerSoftware Users Guide  
8-11  
             
Chapter 8 Viewing Voyager PSD Data  
8.3.2 Calibrating  
This section includes:  
Calibrating  
Matching peaks automatically  
Selecting peaks manually  
Solving and plotting  
Applying new constants to the data file  
Selecting calibration peaks for optimum mass accuracy  
Calibrating To calibrate a PSD spectrum:  
1. Click the Spectrum window to activate it. Select the  
spectrum of interest.  
NOTE: You cannot calibrate a result spectrum or a  
composite spectrum you have accessed from the  
Processing History command on the Display menu. For  
more information, see Section 2.4.7, Recalling and  
Rearranging Traces (Processing History).  
8
2. From the Peaks menu, select Peak Label, then select the  
Mass Label Type (peak apex or peak centroid) to use for  
calibration. Click OK.  
3. From the Process menu, select Mass Calibration, then  
select PSD Calibration.  
The PSD Processing dialog box is displayed with the  
Calibration tab selected (Figure 8-5).  
Hint: You can also access this dialog box by selecting  
PSD Processing from the Process menu, then clicking  
the Calibration tab.  
8-12  
Applied Biosystems  
     
Calibrating a PSD Spectrum  
Column  
header  
buttons  
Figure 8-5 PSD Processing Dialog Box with  
Calibration Tab Displayed  
4. Select a PSD Calibration Reference File that you  
generated as described in Section 8.3.4, Creating PSD  
Calibration Reference (.REF) Files. A calibration  
reference file called Angiotensin_Fragments.REF is  
provided with the software.  
NOTE: Use a calibration reference (.REF) file that  
specifies the peak type for reference masses as  
Resolved Isotope Mass (even if they are not resolved  
isotopes). The calibration routine checks peak width to  
Average Masses in the calibration reference file, the  
software mistakes these narrow peaks as isotopically  
resolved and ignores the reference mass. For more  
information on creating calibration reference files, see  
Section 5.3.3, Creating or Modifying a  
Calibration Reference File (.REF).  
Data ExplorerSoftware Users Guide  
8-13  
       
Chapter 8 Viewing Voyager PSD Data  
NOTE: If the calibration reference file is stored on a  
network drive, an error message may display when you  
select the calibration file when performing a calibration.  
If an error message is displayed, copy the file to a local  
drive on your computer using Windows NT Explorer.  
5. Enter Reference Matching Criteria:  
Minimum IntensityPeaks must be above this  
intensity to be considered a match. Select the unit  
for Minimum Intensity, % Relative Intensity or  
% Relative Area.  
Mass TolerancePeaks must be within this  
tolerance of the theoretical mass to be considered  
a match. Select the unit, m/z or ppm.  
6. Select the Peak Weighting Factor. If the calibration  
includes more than two points, you can apply the following  
weighting factors to fit points to the curve:  
8
NoneAll peaks weighted equally  
Inverse Peak WidthNarrower peaks are  
weighted more than broader peaks  
7. Match observed peaks in the spectrum with reference  
masses in the calibration reference file using one of the  
following procedures:  
• “Matching peaks automaticallydescribed below  
• “Selecting peaks manuallyon page 8-16  
8-14  
Applied Biosystems  
     
Calibrating a PSD Spectrum  
Matching peaks If you want the software to compare observed masses in all  
segment spectra included in the .DAT file to reference masses  
in the selected calibration reference file:  
automatically  
1. Click Match.  
If a mass within the tolerance of any of the masses  
listed in the calibration reference file is found in any  
spectrum in the .DAT file, the match is displayed in the  
Calibration Mass Peak Selection window.  
CAUTION  
Use the Match function with care. Before using, ensure  
that all peaks in all segments are correctly peak  
detected. The Match function examines the complete  
mass range in all segments in the .DAT file.  
NOTE: The entire mass range of each spectrum in the  
.DAT file is checked, not just the mass range included  
in the composite spectrum, and even if the spectrum is  
not currently displayed in the Spectrum window.  
8
For comparison, the difference between the reference  
mass in the calibration reference file and the observed  
peak mass is displayed.  
Hint: You can sort the list of matches by clicking on the  
column header buttons (see Figure 8-5 on page 8-13).  
You can display complete information about a reference  
mass by double-clicking on the mass.  
If no matches are found, an error message is displayed.  
2. To delete a match from the list, select the mass and click  
Delete Selected Match.  
To clear the entire list, click Delete Entire List.  
3. Complete the calibration as described in Solving and  
plottingon page 8-18.  
Data ExplorerSoftware Users Guide  
8-15  
         
Chapter 8 Viewing Voyager PSD Data  
Selecting peaks For optimum mass accuracy, select peaks as described in  
Selecting calibration peaks for optimum mass accuracyon  
page 8-19.  
manually  
To manually select the reference mass for a peak:  
1. Right-click-drag over the peak of interest.  
The Reference Mass dialog box (Figure 8-6) is  
displayed and lists all masses in the calibration  
reference file. The entry highlighted is the nearest  
match in the calibration reference file to the selected  
peak that is within the mass tolerance specified.  
8
Figure 8-6 Reference Mass Dialog Box  
8-16  
Applied Biosystems  
           
Calibrating a PSD Spectrum  
2.  
Do any of the following:  
Click OK to accept the highlighted reference mass  
for matching.  
Select a different reference mass and click OK.  
Type new reference mass information in the  
Name, Theoretical Mass, Charge and Elemental  
Composition text boxes and select the mass type.  
Click OK to accept the reference mass for  
matching.  
NOTE: You must type in a minus sign (-) for  
negative charge states.  
Type new reference mass information in the  
add the information to the calibration reference  
file, then click OK to accept the reference mass for  
matching.  
8
The PSD Calibration dialog box is displayed again (see  
Figure 8-5 on page 8-13) with the observed mass and  
the reference mass you selected displayed in the  
Matched Peak list.  
3.  
Repeat step 1 through step 2 until the desired peaks  
are in the matched list.  
For optimum mass accuracy, take at least one mass  
from each segment, if possible.  
The software does not allow you to add duplicate  
masses collected with the same PSD Mirror Ratio.  
specified in Reference Matching Criteria, but has the  
same PSD Mirror Ratio, the software allows you to  
add it.  
4. Complete the calibration as described in Solving and  
plottingon page 8-18.  
Data ExplorerSoftware Users Guide  
8-17  
 
Chapter 8 Viewing Voyager PSD Data  
Solving and After matching peaks, click Solve and Plot.  
plotting  
The calibration statistics are displayed in the Result tab of the  
Output window and the calibration constants are applied to  
the spectrum displayed  
If you calibrate more than one time, subsequent calibration  
statistics are added to the end of the list in the Output window.  
Older calibration statistics are listed at the top of the list. Use  
the scroll bar to view newer statistics at the bottom of the list.  
The spectrum is calibrated and displayed with an MC trace  
label. The calibration statistics are displayed in the Output  
window.  
Applying new To apply the calibration constants to each spectrum (all  
segments) in the data file, click Apply Calibration.  
constants to the  
data file  
All spectra in the data file are calibrated and displayed with an  
MC trace label. The calibration constants are saved with the  
data file. Each spectrum in the data file is calibrated when  
displayed.  
8
The software uses different values in the calibration equation  
calculation based on the type of calibration performed.  
8-18  
Applied Biosystems  
         
Calibrating a PSD Spectrum  
Selecting To improve calibration statistics, you can select the same  
fragment ion from more than one segment. Monoamino acid  
calibration peaks  
for optimum mass  
accuracy  
fragments (immonium ions) below 150 Da are useful for this  
purpose. Because the segments have been collected with  
different PSD Mirror Ratios, the software allows you to add the  
same mass to the list multiple times.  
Hint: A calibration reference file called  
Immonium_Ions.REF is provided with the software.  
To optimize mass accuracy across the entire PSD composite  
spectrum (multi-point calibration):  
Select reference peaks from a wide range of segments to  
ensure that high and low Mirror Ratios are represented.  
Include peaks with masses that are substantially below  
the Max Stitch Mass (for example, up to 50 percent lower  
than the Max Stitch Mass.  
8
Avoid selecting peaks with masses that are above the  
Max Stitch Mass.  
Avoid selecting peaks with signal-to-noise ratios less  
than 20.  
Data ExplorerSoftware Users Guide  
8-19  
     
Chapter 8 Viewing Voyager PSD Data  
8.3.3 Creating PSD Calibration (.CAL) Files  
and Applying to Other Data Files  
Creating PSD To generate a PSD .CAL file:  
.CAL files  
1. Acquire a standard, for example, angiotensin, in the  
Instrument Control Panel in PSD mode. For more  
information, see the Voyager Biospectrometry  
Workstation Users Guide.  
2. Open the .DAT file in the Data Explorer software.  
3. Calibrate as described in Section 8.3.2, Calibrating.  
4. From the File menu, select Export, then select  
Calibration.  
5. Name and save the file.  
8
Hint: Include a _PSD suffix when you export a PSD  
.CAL file to help you distinguish them from non-PSD  
.CAL files. For example, type Cal_PSD as the file  
name. The complete file name will be Cal_PSD.CAL.  
Applying new You can apply calibration constants from a .CAL file to any  
data file. To apply the new constants from a mass calibration  
file to a different file:  
constants to  
additional files  
1. Display the spectrum to calibrate.  
2. From the Process menu, select Mass Calibration.  
3. Select Import Calibration.  
4. Select the .CAL file to use. Click Open.  
The software displays the calibrated spectrum with an  
MC trace label.  
5. To save the calibration to the data file, select Mass  
Calibration from the Process menu, then select Apply  
Calibration.  
8-20  
Applied Biosystems  
         
Calibrating a PSD Spectrum  
8.3.4 Creating PSD  
Overview You can manually create a calibration reference file by typing  
masses in a text file as described in Section 5.3.3, Creating or  
Modifying a Calibration Reference File (.REF).  
You can also use the Ion Fragmentation calculator to generate  
theoretical fragments and masses from the sequence for a  
standard compound, then automatically save the masses and  
associated information in a calibration reference file. For  
detailed information on using the Ion Fragmentation calculator,  
see Section 6.5, Using the Ion Fragmentation Calculator.  
Using the Ion To create a PSD calibration reference file using the Ion  
Fragmentation calculator:  
Fragmentation  
calculator to  
create a  
calibration  
reference (.REF)  
file  
1. From the Applications menu, select Ion Fragmentation  
Calculator.  
8
(Figure 8-4 on page 8-8).  
2. In the Sequence text box, type the amino acid sequence  
of the standard compound. Use single-letter codes. Set  
other parameters as needed. For parameter descriptions,  
see Section 6.5, Using the Ion Fragmentation Calculator.  
3. Select Monoisotopic from the Calculate fragment ion  
masses as drop-down list.  
Data ExplorerSoftware Users Guide  
8-21  
         
Chapter 8 Viewing Voyager PSD Data  
NOTE: This selection determines the mass type specified  
for the reference masses in the calibration reference file.  
Use a calibration reference (.REF) file that specifies the  
peak type for reference masses as Resolved Isotope  
Mass (even if they are not resolved isotopes). The  
calibration routine checks peak width to determine if a  
Masses in the calibration reference file, the software  
mistakes these narrow peaks as isotopically resolved  
and ignores the reference mass. For more information  
on creating calibration reference files, see  
Section 5.3.3, Creating or Modifying a  
Calibration Reference File (.REF).  
4. Click Induce Fragmentation.  
Results are listed in the ions table, and fragment ions  
are labeled if they are present in the spectrum.  
8
5. Click Create Reference File.  
A Save As dialog box is displayed.  
6. Name and save the calibration reference file.  
Hint: PSD calibration reference files are named with an  
.REF extension. When you create PSD calibration  
reference files, include a _PSD suffix when you name  
files to help you distinguish them from non-PSD  
calibration reference files.  
Hint: Calibration reference files are ASCII text files.  
You can delete unwanted fragment ion entries using  
Microsoft Notepad.  
8-22  
Applied Biosystems  
Calibrating a PSD Spectrum  
8.3.5 Changing the Precursor Mass  
When to change When analyzing the composite spectrum, you may find that  
the observed fragments and sequence are not consistent with  
precursor mass  
the precursor mass used to acquire the .DAT file. For  
example, you acquired the data with a precursor mass of  
1,000.5 Da, and while examining the data in Data Explorer,  
you realize that the data may correspond to a precursor mass  
of 1,000 Da or 1,001 Da.  
You can change the precursor mass in the Data Explorer  
software and regenerate the composite spectrum to observe  
whether the mass accuracy of the fragment ions improves with  
a different precursor mass specified.  
Before changing the precursor mass, note the following points:  
The Change Mass function changes the value for mp in  
the PSD calibration equation (described on page 8-6). It  
does not change the value for tp.  
8
Use the Change Mass function only if the precursor mass  
used to acquire the data does not correspond to the  
sequence you are observing in Data Explorer.  
Do not use the Change Mass function to correct a  
precursor mass that you incorrectly typed before  
acquiring. The precursor ion flight time associated with  
the incorrect mass is stored in the .DAT file and will yield  
invalid fragment masses even if you correct the precursor  
mass. You must specify the correct precursor mass and  
reacquire the data to obtain correct mass and flight time  
values to yield a valid fragment mass calibration.  
Data ExplorerSoftware Users Guide  
8-23  
         
Chapter 8 Viewing Voyager PSD Data  
Changing If the precursor mass taken from the data file is not correct:  
1. Display the Segments tab (see Figure 8-2 on page 8-4)  
by doing either of the following:  
In the PSD Calibration dialog box, click the  
Segments tab  
From the Process menu, select PSD Processing  
2. Click Change Mass and enter a new mass to use for  
calibration.  
3. Click OK.  
4. Click Plot to display the composite spectrum for the new  
mass. The following occurs:  
A new composite spectrum is generated as  
described in How the composite spectrum is  
generatedon page 8-6, and displayed.  
8
The PSD calibration for the data file is updated  
with the new precursor ion mass.  
The new composite spectrum is displayed.  
NOTE: The Max Stitch Masses displayed in the  
PSD segment list in the PSD Processing dialog  
box are not updated until you apply the new  
calibration.  
5. To save the updated calibration information in the data file,  
click Apply Precursor Mass.  
8-24  
Applied Biosystems  
 
Chapter  
9
9 Troubleshooting  
This chapter contains the following sections:  
9.1  
9.2  
9.3  
Overview ......................................................... 9-2  
General Troubleshooting.................................. 9-3  
Processing, Tools, and Applications  
Troubleshooting............................................... 9-6  
9.4  
9.5  
9.6  
Calibration Troubleshooting ........................... 9-10  
Printing Troubleshooting ................................ 9-14  
Peak Detection and Labeling  
Troubleshooting............................................. 9-15  
Data ExplorerSoftware Users Guide  
9-1  
   
Chapter 9 Troubleshooting  
9.1 Overview  
This section includes:  
General troubleshooting  
Processing, tools, and applications troubleshooting  
Calibration troubleshooting  
Printing troubleshooting  
Peak detection and labeling troubleshooting  
Troubleshooting information is organized according to  
likelihood of possible cause, from most likely to least likely  
possible cause. If you are unable to solve your problem using  
the information in the following tables, call Applied Biosystems  
Technical Support. To reach Applied Biosystems Technical  
Support, refer to the list of offices on the back cover of this  
book.  
9
9-2  
Applied Biosystems  
       
General Troubleshooting  
9.2 General Troubleshooting  
Table 9-1 General TroubleshootingMariner and Voyager  
Symptom  
Possible Cause  
Action  
Cannot find data file  
Did not save the spectrum Reacquire spectrum. Save  
to a .DAT file  
.DAT file.  
Error message displayed  
when opening PSD data  
file  
You are currently  
Stop the Experiment. See  
the Voyager  
Biospectrometry  
acquiring the PSD DAT  
file and have not stopped  
the experiment  
Workstation Users Guide.  
Parts of other software  
windows are displayed on windows overlap  
top of the Data Explorer  
Multiple application  
Close applications not in  
use, then minimize and  
maximize the Data  
window, or toolbar buttons  
or status indicators are  
Explorer window to  
refresh the display.  
not displayed  
Failed to create empty  
Chromatogram and  
Resize the Output window  
documentmessage  
Spectrum windows are too (click-drag border) so that  
displayed when you open small  
a data file  
at least a small part of the  
Chromatogram or  
Spectrum window is  
displayed.  
9
Graphic Option settings  
are not applied to all  
traces  
Did not select Use same  
settings for all graphs  
before entering settings  
Select Use same  
settings for all graphs  
then click Apply.  
Data ExplorerSoftware Users Guide  
9-3  
         
Chapter 9 Troubleshooting  
Table 9-1 General TroubleshootingMariner and Voyager (Continued)  
Symptom  
Possible Cause  
Action  
M/z range in data files  
converted to centroid  
M/z range in a data file  
that is converted from  
No action. Normal  
occurrence.  
does not match m/z range profile to centroid is  
in original data file  
determined by the peak  
detection range set in  
Data Explorer, not the m/z  
range in the original data  
file  
Text annotation from a  
previous trace displayed  
on current trace  
Annotation stays in the  
view until you delete it,  
even if you advance to the Annotating Traces.  
next spectrum  
Delete the annotation.  
See Section 2.4.9,  
Table 9-2 General TroubleshootingMariner Only  
Symptom  
Possible Cause  
Action  
Failed to open  
You are opening a data  
file collected with version  
2.1 software and the  
.CGM file not located in  
the same directory as the  
.SPC file  
Place the .CGM and .SPC  
files in the same directory.  
chromatogram data”  
message displayed when  
you open a data file  
9
9-4  
Applied Biosystems  
   
General Troubleshooting  
Table 9-2 General TroubleshootingMariner Only (Continued)  
Symptom  
Possible Cause  
You are viewing  
Action  
Spectra labeled with  
No action. Normal  
occurrence.  
spectrum numbers that do event-filtered MS Method  
not correspond to the axis data. Spectra in an  
event-filtered trace are  
numbered contiguously  
(1,2,3...) regardless of  
their relation to the overall  
acquisition. However, the  
axis of the trace reflects  
the numbering of the  
overall experiment.  
Conversion of .SPC to  
.DAT failed  
A .DAT file with the same  
file name you are  
specifying for the  
converted .SPC file is  
open  
1. Close the open .DAT  
file.  
2. Use a different name  
for the converted .SPC  
file to prevent the new  
conversion from  
overwriting the existing  
file.  
9
Table 9-3 General TroubleshootingVoyager Only  
Symptom  
Possible Cause  
Action  
Failed to get original  
calibrationmessage  
You are opening a  
Voyager .MSF data file  
No action. Normal  
occurrence.  
displayed when you open collected with version 4.0  
a data file or earlier software.  
Data ExplorerSoftware Users Guide  
9-5  
     
Chapter 9 Troubleshooting  
9.3 Processing, Tools, and  
Applications Troubleshooting  
Table 9-4 Processing, Tools, and Applications TroubleshootingMariner and  
Voyager  
Symptom  
Possible Cause  
Failed to calculate result  
for isotope calculator  
You may have tried to  
remove a group that is not  
present in the formula  
Can only remove a group  
that is present in the  
formula. For information,  
see Section 6.2, Using the  
Isotope Calculator.  
All traces in an overlaid  
trace are not processed  
Only the active trace in an Display individual traces,  
overlaid trace is  
processed  
select Use same settings  
for all graphs from  
Graphic Options, then  
process.  
For more information, see:  
Section 1.5, Setting  
Graphic Options  
9
Section 2.4.8,  
Overlaying Traces  
9-6  
Applied Biosystems  
       
Processing, Tools, and Applications Troubleshooting  
Table 9-4 Processing, Tools, and Applications TroubleshootingMariner and  
Voyager (Continued)  
Symptom  
Possible Cause  
Action  
Results not saved for all  
Only results for the active Display individual traces,  
traces in an overlaid trace trace are saved  
then save results for each  
trace.  
For more information, see:  
Section 2.6, Saving,  
Deleting .DAT  
Results  
Section 2.8, Saving,  
Opening, and  
Deleting .SPC  
Results Files  
(Mariner Data Only)  
Link View command does You did not select Link  
You must select Link View  
for each window and each  
data file you want to link.  
See Linking viewson  
page 2-13.  
not have an effect on  
View for each window or  
windows or open data files data file  
9
Data ExplorerSoftware Users Guide  
9-7  
 
Chapter 9 Troubleshooting  
Table 9-4 Processing, Tools, and Applications TroubleshootingMariner and  
Voyager (Continued)  
Symptom  
Possible Cause  
After Single-charge  
Conversion of multiply  
charged peaks, you see  
charge states other than 0  
or 1  
Peaks in the original  
spectrum are labeled with  
an incorrect charge state  
1. Set peak detection  
thresholds to disregard  
these peaks. See  
Section 3.2.5, Charge  
and Examples.  
2. Convert the spectrum  
again. See  
Section 5.10,  
Converting to a Singly  
Charged Spectrum  
(Mariner Data Only).  
9
9-8  
Applied Biosystems  
Processing, Tools, and Applications Troubleshooting  
Table 9-5 Processing, Tools, and Applications TroubleshootingMariner Only  
Symptom  
Possible Cause  
Failed to calculate result  
for mass deconvolution  
Did not select at least two Select at least two peaks  
peaks for same charge  
envelope  
(for example, +1 and +2).  
For more information, see  
Section 5.6, Mass  
Deconvolution  
(Mariner Data Only).  
Multiple Charge/Mass  
Your system does not  
Contact Applied  
Deconvolution commands include the optional mass Biosystems to purchase  
dimmed on Process menu deconvolution software  
the option.  
Centroiding, Mass  
Calibration, Multiple  
Charge commands not  
displayed on Process  
menu  
Spectrum window not  
active  
Activate Spectrum  
window, then select  
Process menu.  
Resolution command not  
displayed on Tools menu  
Spectrum window not  
active  
Activate Spectrum  
window, then select Tools  
menu.  
Only the active trace  
zooms in Overlay mode  
Zooming may not behave  
as expected on overlaid  
traces  
Click  
in the top right  
9
corner of the window to  
restore the view.  
Data ExplorerSoftware Users Guide  
9-9  
       
Chapter 9 Troubleshooting  
9.4 Calibration Troubleshooting  
Table 9-6 Calibration TroubleshootingMariner and Voyager  
Symptom  
Possible Cause  
Auto Calibration is turned Current spectrum is not  
Advance to the next  
spectrum, then return to  
the current spectrum. For  
more information, see  
Section 5.4, Automatic  
Calibration.  
on, but current spectrum  
is not auto calibrated  
calibrated until the next  
time the spectrum is  
displayed  
During calibration, the  
software is not matching  
spectrum masses to some determine if a peak  
reference masses in the  
calibration reference file  
The calibration routine  
checks peak width to  
When specifying highly  
charged, non-isotopically  
resolved species with  
peaks less than 1 Da wide  
(for example, myoglobin  
+20) as reference masses  
in a calibration reference  
Resolved Isotope Mass  
(even though it is not a  
resolved isotope). See  
Modifying a  
matches a Resolved  
Isotope Mass or an  
Average mass. If narrow  
peaks are specified as  
Average Masses in the  
calibration reference file,  
the software mistakes  
these narrow peaks as  
isotopically resolved and  
ignores the reference  
mass.  
9
Calibration Reference File  
(.REF).  
Mass tolerance or  
Adjust. See Section 5.3,  
Minimum Intensity set too Manual Calibration.  
high  
9-10  
Applied Biosystems  
     
Calibration Troubleshooting  
Table 9-6 Calibration TroubleshootingMariner and Voyager (Continued)  
Symptom  
Possible Cause  
Action  
Calibration returns an  
invalid number of matches mass list in Manual or  
When creating a reference Do not include multiple  
entries with the same m/z  
Automatic calibration, the value in the Calibration  
software allows you to add Reference file. See  
multiple items with the  
same m/z value to the  
calibration list box if any  
other attribute of the  
reference compound is  
different (for example,  
charge state or name).  
Each mass in the list is  
considered during  
Section 5.3.3, Creating or  
Modifying a  
Calibration Reference File  
(.REF).  
calibration.  
.REF file created with  
Windows Notepad not  
listed when you select  
Reference file  
Reference file does not  
include .REF extension  
Some applications  
automatically append a  
.TXT extension to file  
names.  
To name the file with a  
.REF extension, include  
the file name and  
9
extension in  
double-quotes in the Save  
File dialog box, for  
example CAL.REF.  
Data ExplorerSoftware Users Guide  
9-11  
   
Chapter 9 Troubleshooting  
Table 9-7 Calibration TroubleshootingMariner Only  
Symptom  
Possible Cause  
Action  
Mass Calibration  
commands are dimmed  
Chromatogram window is  
selected  
Select Spectrum window.  
Apply Calibration  
command is dimmed  
when calibrating  
MS Method data  
For MS Method data,  
calibration is valid for an  
individual spectrum. You  
cannot apply the  
No action. Normal  
occurrence.  
calibration from one  
spectrum to the entire  
data file.  
Error displayed when you .CAL file corrupted  
import a calibration  
Create new .CAL file. See  
Exporting .BIC, .MSM,  
and .CAL fileson  
page 1-36.  
Importing a .CAL file  
generated from a Voyager  
data file  
Import a Mariner .CAL.  
9
9-12  
Applied Biosystems  
   
Calibration Troubleshooting  
Table 9-8 Calibration TroubleshootingVoyager Only  
Symptom  
Possible Cause  
Action  
Error displayed when you .CAL file corrupted  
import a calibration  
Create new .CAL file. See  
Exporting .BIC, .MSM,  
and .CAL fileson  
page 1-36.  
Importing a .CAL file  
generated from a Mariner  
data file  
Import a Voyager .CAL.  
Importing a .CAL file  
Import a .CAL generated  
generated from a data file from a data file collected  
collected in a different  
instrument mode (Linear,  
Reflector, or PSD)  
in the same instrument  
mode.  
Importing a .CAL file  
generated on a different  
instrument  
Import a .CAL generated  
on the same instrument.  
9
Data ExplorerSoftware Users Guide  
9-13  
 
Chapter 9 Troubleshooting  
9.5 Printing Troubleshooting  
Table 9-9 Printing TroubleshootingMariner and Voyager  
Symptom  
Possible Cause  
Action  
Traces do not print  
Line width is set to 0 or 1  
See Section 1.5, Setting  
Graphic Options.  
Line color is set to white  
Change the color. See  
Section 1.5, Setting  
Graphic Options.  
Landscape printer setup  
lost when you close Data  
Explorer  
Landscape option set  
using Printer Setup in  
Data Explorer  
Set landscape printing  
using Printer Settings in  
Windows Control Panel.  
See Dedicating a printer  
to landscape orientation”  
on page 2-35.  
Data file names do not  
Some printers may not  
open only two data files  
before you select Print All  
Views. For more  
print for multiple data files print the data file name if  
you select Print All Views  
from the File menu with  
more than two data files  
open. We have observed  
this on HP LaserJet4,  
information, see  
Section 2.4.11, Printing  
Traces.  
9
HP LaserJet5, and  
HP LaserJet6 printers.  
9-14  
Applied Biosystems  
       
Peak Detection and Labeling Troubleshooting  
9.6 Peak Detection and  
Labeling Troubleshooting  
This section includes:  
Peak detection and labeling troubleshooting  
Charge state and isotope determination troubleshooting  
Table 9-10 Peak Detection and Labeling TroubleshootingMariner and  
Voyager  
Symptom  
Possible Cause  
Action  
Peaks are not detected or Peaks are very close  
Zoom in on region of  
interest.  
labeled  
together or label is too  
long  
Select Allow  
overlapping peak  
labels in the Peak  
Label dialog box. See  
and Spectrum Peak  
Labels.  
Peak detection  
Adjust parameters. See  
Section 3.2, Peak  
Detection.  
parameters, in particular,  
Filter Width, not set to  
detect peaks  
Peak filtering is enabled.  
Only peaks that meet the  
peak list filtering criteria  
are labeled.  
Disable peak list filtering.  
See Filtering  
the spectrum peak liston  
page 3-42.  
Data ExplorerSoftware Users Guide  
9-15  
     
Chapter 9 Troubleshooting  
Table 9-10 Peak Detection and Labeling TroubleshootingMariner and Voyager  
(Continued)  
Symptom  
Possible Cause  
Peaks are not detected or Analyzing masses above  
Increase Mass Resolution  
Strategy for Voyager Peak  
Detection.  
labeled  
20,000 Da  
(continued)  
%Max Peak Area set too  
high  
Decrease. See  
Section 3.2.2, Strategy for  
%Base Peak Intensity set Set to 0. Adjust %Max  
too high  
Peak Area to optimize  
peak detection for Voyager  
data. See Section 3.2.2,  
Strategy for Voyager Peak  
Detection.  
When creating a custom  
label for a spectrum, you  
right-click-drag across a  
peak to identify the peak,  
and an extracted ion  
chromatogram is created  
instead  
The Chromatogram  
Make sure the Spectrum  
window is active before  
selecting Peak Label from  
the Display menu.  
window was active when  
you selected Peak Label  
from the Display menu  
9
9-16  
Applied Biosystems  
   
Peak Detection and Labeling Troubleshooting  
Table 9-10 Peak Detection and Labeling TroubleshootingMariner and Voyager  
(Continued)  
Symptom  
Possible Cause  
Action  
Expected user label not  
displayed  
Delta X value includes  
more than one peak apex enough to prevent the  
peak labeling windows  
Set Delta X value low  
from overlapping. For  
information, see  
Section 3.5.3, Setting  
Custom Peak Labels.  
Delta X value is outside  
acquisition range for the  
data file  
Set Delta X value within  
acquisition range for the  
data file. For information,  
see Section 3.5.3, Setting  
Custom Peak Labels.  
Peak label placed on peak Filter Width Increment set Set Increment to 1. See  
shoulder instead of peak  
apex  
higher than 1  
Incrementon page 3-31.  
Table 9-11 Peak Detection and Labeling TroubleshootingVoyager Only  
Symptom  
Possible Cause  
Action  
Section 3.2.2, Strategy for  
Voyager Peak Detection.  
Noise detected as peaks  
%Max Peak Area set too  
low  
Low Mass Gate spike  
identified as Base Peak  
(linear data)  
Software identifies the  
most intense peak in a  
trace as the base peak,  
and does not ignore  
artifacts  
Truncate the data. See  
Section 5.9, Truncating a  
Spectrum.  
Data ExplorerSoftware Users Guide  
9-17  
Chapter 9 Troubleshooting  
Table 9-11 Peak Detection and Labeling TroubleshootingVoyager Only  
Symptom  
Possible Cause  
Action  
Partially resolved peaks  
not detected  
Mass resolution set too  
high to detect average  
mass  
Decrease Mass  
Section 3.2.2, Strategy for  
Voyager Peak Detection.  
%Base Peak Intensity not Adjust. See Section 3.2.2,  
adjusted correctly  
Strategy for Voyager Peak  
Detection.  
Table 9-12 Charge State and Isotope Determination TroubleshootingMariner  
Only  
Symptom  
Known isotope labeled  
with incorrect charge state parameter set too low.  
(too low)  
Max Charge State  
Set Max Charge State  
correctly. See Peak  
Processing parameters  
(spectrum data only)on  
page 3-26.  
See example in Max  
Charge State set too low”  
on page 3-33.  
9
9-18  
Applied Biosystems  
       
Peak Detection and Labeling Troubleshooting  
Table 9-12 Charge State and Isotope Determination TroubleshootingMariner  
Only (Continued)  
Symptom  
Possible Cause  
Action  
Known isotope not labeled Charge State peak labels  
Turn on Charge State  
peak labels. See  
Section 3.5.2, Setting  
with charge state  
disabled  
Labels.  
Max Charge State  
Set Max Charge State  
correctly. See Peak  
page 3-26.  
parameter set too low.  
See example in Max  
Charge State set too low”  
on page 3-33.  
Peak List Filtering is  
enabled, with Charge  
State filter enabled and  
set too low  
Disable Peak List  
Minimum Intensity set too Set Minimum Intensity  
high to detect other  
isotope peaks. See  
example in Effect of  
Minimum Intensityon  
page 3-36.  
correctly. See Advanced  
Settings (spectrum data  
only)on page 3-28.  
Filter width set too high to Set Filter Width correctly.  
determine other isotope  
peaks  
See Advanced Settings  
(spectrum data only)on  
page 3-28.  
Data ExplorerSoftware Users Guide  
9-19  
Chapter 9 Troubleshooting  
Table 9-12 Charge State and Isotope Determination TroubleshootingMariner  
Only (Continued)  
Symptom  
Possible Cause  
Action  
Spectrum peaks not  
labeled with charge state  
when charge state labels  
are selected  
Mass of original molecule No action. Normal  
above 4,000 Da, not  
range in which the Mariner  
system can resolve the  
isotope peaks  
occurrence.  
Charge State parameters  
not set to detect at least  
two isotope peaks  
Adjust parameters. See  
Peak Processing  
parameters (spectrum  
data only)on page 3-26.  
Peaks are more than 1 Da No action. Normal  
apart  
occurrence.  
Peaks are not from the  
same isotope species  
No action. Normal  
occurrence.  
Filter width is set too high Adjust Filter Width. See  
to detect other isotope  
peaks  
Advanced Settings  
page 3-28.  
Maximum Charge State  
for charge state  
determination is set lower Processing parameters  
Adjust the Maximum  
Charge State. See Peak  
9
than the charge state of  
the peak  
(spectrum data only)on  
page 3-26.  
9-20  
Applied Biosystems  
     
Peak Detection and Labeling Troubleshooting  
Table 9-12 Charge State and Isotope Determination TroubleshootingMariner  
Only (Continued)  
Symptom  
Possible Cause  
Charge state  
Action  
Spectrum peaks labeled  
Adjust parameters. See:  
with incorrect charge state determination parameters  
parameters  
on page 3-26  
when charge state labels  
are selected  
are set such that peaks  
are determined to have no  
charge  
(continued)  
Section 3.2.5,  
Charge State  
Noise between isotope  
peaks is detected  
Increase %Base Peak  
Intensity or %Max Peak  
Area to eliminate noise.  
See Basic Settings  
(spectrum data only)on  
page 3-22.  
Apply noise filter. See  
Section 5.7, Noise  
Filtering/Smoothing.  
Monoisotopic peak not  
labeled correctly  
Peak detection charge  
state parameters are not  
set correctly, the software monoisotopic peak. See  
Adjust parameters to  
correctly identify the  
is identifying the tallest  
peak in a cluster as the  
monoisotopic peak  
Peak Processing  
parameters (spectrum  
data only)on page 3-26.  
Data ExplorerSoftware Users Guide  
9-21  
   
Chapter 9 Troubleshooting  
9
9-22  
Applied Biosystems  
Appendix  
A
A Warranty  
Applied Biosystems supplies or recommends certain  
configurations of computer hardware, software, and  
peripherals for use with its instrumentation. Applied  
Biosystems reserves the right to decline support for or  
impose extra charges for supporting non-standard  
computer configurations or components that have not  
been supplied or recommended by Applied Biosystems.  
Applied Biosystems also reserves the right to require that  
computer hardware and software be restored to the  
standard configuration prior to providing service or  
technical support.  
Limited Product Warranty  
Limited warranty Applied Biosystems warrants that for a period of ninety (90)  
days from the date of installation, the Data Explorersoftware  
designated for use with MarinerAPI-TOF Workstations or  
VoyagerBiospectrometryWorkstations will perform  
substantially in accordance with the function and features  
described in its accompanying documentation when properly  
installed on the instrument system. Applied Biosystems does  
Data ExplorerSoftware Users Guide  
A-1  
       
Appendix A Warranty  
not warrant that the operation of the instrument or software will  
be uninterrupted or error free. Applied Biosystems will provide  
any software corrections or bug-fixes, if and when they  
become available, for a period of ninety (90) days after  
installation.  
A
Warranty period Any applicable warranty period under these sections will begin  
on the date of installation, but no later than three (3) months  
effective date  
from the date of shipment, for software installed by Applied  
Biosystems personnel, unless that date has been delayed at  
the buyers request, but in no event later than thirty (30) days  
after shipment. In that case, and for all hardware and software  
installed by the buyer, and for all other products, the applicable  
warranty period begins the date the component is delivered to  
the buyer. For software installed by the buyer or any person  
other than Applied Biosystems, the applicable warranty period  
will begin the date the product is delivered to the buyer.  
Warranty claims Warranty claims must be made within the applicable warranty  
period or, for chemicals or other consumable products, within  
thirty (30) days after receipt by the buyer.  
Warranty The above warranties do not apply to defects resulting from  
misuse, neglect, or accident, including without limitation:  
exceptions  
operation outside of the environmental or use specifications or  
not in conformance with the instructions for the instrument  
system, software, or accessories; performance of improper or  
inadequate maintenance by the user; installation of software  
or interfacing not supplied by Applied Biosystems; and  
modification or repair of the instrument or the software not  
authorized by Applied Biosystems.  
THE FOREGOING PROVISIONS SET FORTH APPLIED  
BIOSYSTEMS SOLE AND EXCLUSIVE  
REPRESENTATIONS, WARRANTIES, AND OBLIGATIONS  
WITH RESPECT TO ITS PRODUCTS, AND APPLIED  
BIOSYSTEMS MAKES NO OTHER WARRANTY OF ANY  
KIND WHATSOEVER, EXPRESSED OR IMPLIED,  
INCLUDING WITHOUT LIMITATION, WARRANTIES OF  
MERCHANTABILITY AND FITNESS FOR A PARTICULAR  
PURPOSE, WHETHER ARISING FROM A STATUTE OR  
OTHERWISE IN LAW OR FROM A COURSE OF DEALING  
OR USAGE OF TRADE, ALL OF WHICH ARE EXPRESSLY  
DISCLAIMED.  
A-2  
Applied Biosystems  
 
Warranties THE REMEDIES PROVIDED HEREIN ARE BUYERS SOLE  
AND EXCLUSIVE REMEDIES. WITHOUT LIMITING THE  
limitations  
A
GENERALITY OF THE FOREGOING, IN NO EVENT SHALL  
APPLIED BIOSYSTEMS BE LIABLE, WHETHER IN  
CONTACT, TORT, WARRANTY OR UNDER ANY STATUTE  
(INCLUDING WITHOUT LIMITATION ANY TRADE  
PRACTICE, UNFAIR COMPETITION OR OTHER STATUTE  
OF SIMILAR IMPORT) OR ON ANY OTHER BASIS, FOR  
DIRECT, INDIRECT, PUNITIVE, INCIDENTAL,  
CONSEQUENTIAL, OR SPECIAL DAMAGES SUSTAINED  
BY BUYER OR ANY OTHER PERSON OR ENTITY,  
WHETHER OR NOT FORESEEABLE AND WHETHER OR  
NOT APPLIED BIOSYSTEMS IS ADVISED OF THE  
POSSIBILITY OF SUCH DAMAGES, INCLUDING WITHOUT  
LIMITATION, DAMAGES ARISING FROM OR RELATED TO  
LOSS OF USE, LOSS OF DATA, FAILURE OR  
INTERRUPTION IN THE OPERATION OF ANY EQUIPMENT  
OR SOFTWARE, DELAY IN REPAIR OR REPLACEMENT,  
OR FOR LOSS OF REVENUE OR PROFITS, LOSS OF  
GOOD WILL, LOSS OF BUSINESS OR OTHER FINANCIAL  
LOSS OR PERSONAL INJURY OR PROPERTY DAMAGE.  
NO AGENT, EMPLOYEE, OR REPRESENTATIVE OF  
APPLIED BIOSYSTEMS HAS ANY AUTHORITY TO BIND  
APPLIED BIOSYSTEMS TO ANY AFFIRMATION,  
REPRESENTATION, OR WARRANTY CONCERNING THE  
PRODUCT THAT IS NOT CONTAINED IN THIS LIMITED  
WARRANTY STATEMENT. ANY SUCH AFFIRMATION,  
REPRESENTATION, OR WARRANTY MADE BY ANY  
AGENT, EMPLOYEE, OR REPRESENTATIVE OF APPLIED  
BIOSYSTEMS WILL NOT BE BINDING ON APPLIED  
BIOSYSTEMS.  
THIS WARRANTY IS LIMITED TO THE BUYER OF THE  
PRODUCT FROM APPLIED BIOSYSTEMS AND IS NOT  
TRANSFERABLE.  
Data ExplorerSoftware Users Guide  
A-3  
Appendix A Warranty  
A
A-4  
Applied Biosystems  
Appendix  
B
B Overview  
of Isotopes  
This appendix contains the following sections:  
B.1 Isotopes .................................................... B-2  
B.2 Monoisotopic and Average Masses ........... B-6  
B.3 Isotopes of Common Elements .................. B-8  
Data ExplorerSoftware Users Guide  
B-1  
     
Appendix B Overview of Isotopes  
B.1 Isotopes  
Overview Many elements in their natural state exist as one of several  
isotopes. An isotope is one of two or more atoms with the  
same atomic number but a different mass. The most abundant  
isotope of carbon is 12C, but natural carbon also contains 13  
and 14C.  
C
Because a mass spectrometer measures mass-to-charge  
ratios, isotopes appear in the mass spectrum. Isotopes of low  
abundance, such as 14C, do not affect the appearance of a  
mass spectrum. However, isotopes that occur in greater  
abundance, such as 13C, which occurs in a natural abundance  
of approximately 1.1 percent1 in carbon, do affect the  
appearance of a mass spectrum.  
B
The mass spectrum of methane (Figure B-1) illustrates the  
impact of an isotope on the appearance of a mass spectrum.  
Methane includes a peak representing the molecular ion at  
16 Da (12CH4) and a peak representing the isotope at 17 Da  
(13CH4). The relative abundance of the ions is about 99:1.  
12C  
13C  
Figure B-1 Mass Spectrum of Methane  
1. Meth. Enzymol., McCloskey, J.A, ed., 1990, 193, 869.  
B-2  
Applied Biosystems  
 
Isotopes  
As the number of carbon atoms in a compound increases, the  
possibility of the compound containing a 13C instead of a 12  
C
also increases. A compound with ten carbon atoms includes a  
molecular ion M+ and an isotopic ion (M+1)+ one mass unit  
greater than the molecular ion, which is approximately  
11 percent of the abundance of the molecular ion. The  
possibility of including two 13C atoms in the same molecule  
also increases with increasing number of carbon atoms.  
Therefore, (M+2)+ ions become more visible. In a compound  
with ten carbon atoms such as decahydro naphthalene,  
C10H18 (Figure B-2), relative heights of M+, (M+1)+, and (M+2)+  
peaks are 100:11:0.5.  
B
M+  
(M+1)+  
(M+2)+  
Figure B-2 Mass Spectrum of Decahydro Naphthalene  
Isotopic pattern All compounds containing carbon include molecular ions and  
isotopic ions that are 1 and 2 mass units higher than the  
molecular ion. At higher masses, the isotopic pattern of a  
mass spectrum is more pronounced as the relative abundance  
of isotopes increases. In angiotensin I (Figure B-3) with a  
molecular weight of 1,296 Da, (C62H89O14N17), a peak  
containing one 13C isotope is approximately 70 percent of the  
pure 12C peak.  
in mass spectra  
Data ExplorerSoftware Users Guide  
B-3  
 
Appendix B Overview of Isotopes  
M+  
(M+1)+  
(M+2)+  
B
Figure B-3 Mass Spectrum of Angiotensin I at  
Resolution 3,000  
In compounds with more than 100 carbon atoms, the height of  
the first 13C isotope peak exceeds the height of the 12C peak.  
Isotope-limited The mass range you analyze and the resolving power of the  
analyzer determine if you observe resolved isotope peaks in a  
mass spectrum.  
resolution  
Figure B-3 represents angiotensin I at a resolution of 3,000.  
Isotopes are fully resolved, and sharp peaks are observed in  
the mass spectrum.  
At lower resolutions, the shape of the individual peaks  
becomes less pronounced, or may not be differentiated at all.  
Figure B-4 represents angiotensin I at a theoretical resolution  
of 1,000.  
B-4  
Applied Biosystems  
   
Isotopes  
M+  
(M+1)+  
(M+2)+  
B
Figure B-4 Mass Spectrum of Angiotensin I at Resolution  
1,000  
If isotopes cannot be resolved, the highest resolution you can  
obtain is limited by the width of the isotopic envelope. The  
isotopic envelope is the mass range of the combined isotopes,  
as measured at the half height of the tallest isotope peak in the  
compound (Figure B-5).  
isotopic envelope  
Figure B-5 Unresolved Isotopes  
Data ExplorerSoftware Users Guide  
B-5  
       
Appendix B Overview of Isotopes  
B.2 Monoisotopic  
and Average Masses  
When isotopes are clearly resolved (Figure B-6), the  
monoisotopic mass is used for mass labeling, and  
corresponds to the lowest mass peak in the cluster.  
Monoisotopic mass  
corresponds to  
lowest mass peak  
B
Figure B-6 Monoisotopic Mass  
When isotopes are not resolved (Figure B-7), the average  
mass is used for mass labeling, and corresponds to the  
centroid of the unresolved peak cluster (weighted average of  
all isotope peaks in the cluster).  
B-6  
Applied Biosystems  
             
Monoisotopic and Average Masses  
Average mass  
corresponds to  
centroid of unresolved  
peak cluster  
B
Figure B-7 Average Mass  
Data ExplorerSoftware Users Guide  
B-7  
 
Appendix B Overview of Isotopes  
B.3 Isotopes of Common Elements  
Table B-1 lists the natural abundance of isotopes for some  
common elements seen in mass spectrometry  
Table B-1 Isotopes of Common Elements1  
Natural  
abundance  
(%)  
Natural  
abundance  
(%)  
Isotope  
Mass  
Isotope  
Mass  
B
1H  
1.0078  
99.985  
0.015  
98.90  
1.10  
31P  
30.9737  
31.9720  
32.9714  
33.9678  
35.9670  
34.9688  
36.9659  
38.9637  
39.9639  
40.9618  
78.9183  
80.9162  
100  
2H  
2.0141  
32S  
33S  
34S  
36S  
35Cl  
37Cl  
39K  
40K  
41K  
79Br  
81Br  
127I  
95.02  
0.75  
12C  
13C  
14N  
15N  
16O  
17O  
18O  
19F  
12  
13.0033  
14.0030  
15.0001  
15.9949  
16.9991  
17.9991  
18.9984  
22.9897  
27.9769  
28.9764  
29.9737  
4.21  
99.63  
0.37  
0.02  
75.77  
24.23  
93.2581  
0.012  
6.7302  
50.69  
49.31  
99.76  
0.04  
0.200  
100  
23Na  
28Si  
29Si  
30Si  
100  
92.23  
4.67  
126.9044  
100  
____  
____  
____  
3.10  
1. Meth. Enzymol., McCloskey, J.A, ed., 1990, 193.  
B-8  
Applied Biosystems  
   
Appendix  
C Data Explorer  
Toolbox  
C
(Visual Basic  
Macros)  
This appendix includes:  
C.1 Overview................................................ C-2  
C.2 Preparing Data  
Before Accessing Macros ....................... C-3  
C.3 Accessing the Macros............................ C-4  
C.4 Using the  
Ladder Sequencing Toolbox.................... C-5  
C.5 Using the  
Peptide Fragmentation Toolbox............... C-9  
C.6 Using the Polymer Analysis Toolbox ..... C-15  
C.7 Using MS Fit/MS Tag Toolbox ............... C-18  
Data ExplorerSoftware Users Guide  
C-1  
   
Appendix C Data Explorer Toolbox (Visual Basic Macros)  
C.1 Overview  
Macros provided The following toolbox of Visual Basic macros is provided with  
the Data Explorer software:  
Ladder Sequencing Use when performing  
sequencing to label peaks with the appropriate amino  
acid, DNA residue, or RNA residue.  
Peptide Fragmentation Use when examining  
Voyager composite PSD spectra or Mariner in-source  
CID spectra to label immonium ions, identify pairs, view  
sequences based on different reference peaks, and  
determine if a selected peak is of a specific fragment  
ion category.  
Polymer Analysis Use when analyzing polymer  
spectra to calculate average molecular weight values.  
MS Fit/MS Tag Use when analyzing peptide or  
C
Importing macros If the macros listed above are not listed when you try to run  
them (described in Section C.3, Accessing the Macros), you  
provided  
must import them into the Data Explorer project. For  
information, see Section 6.7.7, Importing or Exporting Macros  
in DATAEXPLORER.VB6.  
Modifying the The Visual Basic macros described in this document are  
provided in a file called DataExplorer.VB6 in the directory that  
contains the Mariner or Voyager program file. New macros  
you create using the Macro Recorder in the Data Explorer  
software are also added to this file. You can modify the macros  
in this file as desired. However, before you make changes,  
make a copy of the DataExplorer.VB6 file (for example, make  
a copy called DataExplorer.BAK). If the changes you make are  
not acceptable, you can copy or rename the backup file over  
the modified file.  
macros  
C-2  
Applied Biosystems  
                   
Preparing Data Before Accessing Macros  
References The following references are selected by default in the  
DataExplorerProject in the Visual Basic Editor and are  
required for the macros in the DataExplorer.VB6 file to  
successfully run:  
required  
Visual Basic For Applications  
Data Explorer 4.0 Type Library  
OLE Automation  
Microsoft Forms 2.0 Object Library  
Microsoft Internet Controls  
To view references:  
1. Open the DataExplorerProject (select  
VisualBasicEditor from the Tools menu in the Data  
Explorer window).  
2. Select References from the Tools menu in the Visual  
Basic Editor.  
C
C.2 Preparing Data  
Before Accessing Macros  
Before you access the toolbox, open data files of interest and:  
Smooth the data, if necessary, to reduce noise  
Set optimum peak detection and labeling  
Set the display range or zoom as needed  
Resize and organize the displays as needed  
Enable Allow Overlapping Labels (select Peak Label  
from the Peaks menu). If overlapping labels are not  
enabled, some peaks may not be labeled.  
NOTE: Before using the MS Fit/MS Tag toolbox, additional  
data preparation is recommended. See Section Using MS  
Fit/MS Tag Toolbox.  
Data ExplorerSoftware Users Guide  
C-3  
     
Appendix C Data Explorer Toolbox (Visual Basic Macros)  
C.3 Accessing the Macros  
To access the macros:  
1. Open the Data Explorer software.  
2. Open a data file.  
3. Prepare the data as described in the previous section.  
4. From the Tools menu, select Macros.  
5. In the Macros dialog box, select  
modToolBoxPalette.Toolbox_Palette, then click Run.  
The Toolbox Palette dialog box is displayed.  
C
Figure 3-8 Toolbox Palette Dialog Box  
C-4  
Applied Biosystems  
   
Using the Ladder Sequencing Toolbox  
NOTE: If the modToolBoxPalette.Toolbox_Palette is not  
listed, you must import the macro into the Data Explorer  
project. For information, see Section 6.7.7, Importing or  
Exporting Macros in DATAEXPLORER.VB6.  
Hint: You can assign the  
modToolBoxPalette.Toolbox_Palette macro to a macro  
button in the Data Explorer software. For information, see  
the Data Explorer Users Guide, Section 6.7, Using the  
Macro Recorder.  
You can also access individual macros directly, or assign  
individual macros to a macro button, by selecting the  
individual macro in the Macros dialog box.  
C
C.4 Using the  
Ladder Sequencing Toolbox  
Use the Ladder Sequencing toolbox when you perform  
sequencing to label peaks with the appropriate amino acid,  
DNA, or RNA residue. Labels are determined by calculating  
the mass differences between the peaks and comparing the  
values to an internal listing of mass differences and  
corresponding residues.  
Running the To run the Ladder Sequencing macro:  
macro  
1. In the Toolbox Palette dialog box, click Ladder  
Sequencing Toolbox. The range displayed in the  
Spectrum window is reflected in the Mass Range fields.  
NOTE: If you zoom on a different region in the Spectrum  
window, the new range is not updated in the Mass Range  
field, but is used for the analysis when you click Label  
Peaks. The range is updated when you click any button.  
Data ExplorerSoftware Users Guide  
C-5  
     
Appendix C Data Explorer Toolbox (Visual Basic Macros)  
2. Enter the mass tolerance to apply to the analysis.  
3. Click Get Spec Peak List.  
4. Remove peaks you do not want included in the calculation  
by clicking the peak in the list, then clicking Delete  
Selected Peaks/Adducts.  
5. Select the type of spectrum you are examining: Peptide,  
DNA, or RNA.  
6. Under Adducts to Remove, select the items you do not  
want included in the interpretation.  
C
Figure 3-9 Ladder Sequencing Toolbox  
C-6  
Applied Biosystems  
Using the Ladder Sequencing Toolbox  
7. Under Annotate Spectrum, select the types of labels you  
want displayed:  
Reference Mass (*) Mass of the reference  
peak against which the current peak is  
compared.  
Mass Difference Difference between the  
current peak and the reference, preceded by a  
minus sign ().  
Peak Mass Mass of the current peak,  
preceded by an equal sign (=).  
8. Click Label Peaks.  
If Peptide is selected, the software:  
Examines the spectrum in 55 Da increments  
and selects the most intense ion in the range.  
(The 55 Da increment is used because it is  
less than the smallest mass difference related  
to a residue.)  
C
Labels mass differences (plus or minus the  
specified Tolerance) that correspond to amino  
acids.  
Labels the reference peak from which the  
mass difference was derived with an  
asterisk (*).  
Applies the additional labels you selected  
under Annotate Spectrum.  
Data ExplorerSoftware Users Guide  
C-7  
Appendix C Data Explorer Toolbox (Visual Basic Macros)  
If DNA or RNA is selected, the software:  
Examines the spectrum in 270 Da increments  
and selects the most intense ion in the range.  
(The 270 Da increment is used because it is  
less than the smallest mass difference related  
to a DNA or RNA base.)  
Labels mass differences (plus or minus the  
specified Tolerance) that correspond to DNA  
(ACGT) or RNA (ACGU) bases.  
Labels the reference peak from which the  
mass difference was derived with an  
asterisk (*).  
Applies the additional labels you selected  
under Annotate Spectrum.  
Displaying the To display the original labels, select Peak Label from the  
C
Peaks menu in the Data Explorer software, then deselect  
original labels  
User Labels box.  
C-8  
Applied Biosystems  
Using the Peptide Fragmentation Toolbox  
C.5 Using the Peptide  
Fragmentation Toolbox  
Use the Peptide Fragmentation toolbox when examining  
Voyager composite PSD spectra or Mariner in-source CID  
spectra to label immonium ions, identify fragment ion pairs,  
view sequences based on different reference peaks, and  
determine if a selected peak is of a specific fragment ion  
category.  
Setup To enter Setup parameters for the Peptide Fragmentation  
Toolbox macro:  
1. In the Toolbox Palette dialog box, click Peptide  
Fragmentation Toolbox.  
C
Figure 3-10 Peptide Fragmentation Setup  
2. In the Setup tab, click Get Spec Peak List.  
Data ExplorerSoftware Users Guide  
C-9  
     
Appendix C Data Explorer Toolbox (Visual Basic Macros)  
3. Add peaks to the peak list to be included in the calculation  
by typing a mass in the Add Mass to Peak List field, then  
clicking Add Peak.  
Remove unwanted peaks from the list that you do not  
want included in the calculation by clicking the peak in the  
list, then clicking Delete Selected Peaks.  
4. If you will be identifying y and b pairs, select the precursor  
peak, then click Use Selected Peak.  
Pairs To list ion pairs:  
1. Click the Pairs tab.  
C
Figure 3-11 Peptide Fragmentation Pairs  
2. Click the button that corresponds to the ion masses you  
want identified:  
a and b pairs Lists peak pairs with a 28 Da  
mass difference.  
(a, b, y) 17 pairs Lists peak pairs with a 17  
Da mass difference, which corresponds to the  
loss of NH3.  
C-10 Applied Biosystems  
     
Using the Peptide Fragmentation Toolbox  
y and b pairs Lists peak pairs whose  
combined masses plus 1 Da add up to the  
Precursor Ion Mass you specified on the  
Setup tab.  
Loss of H2O Lists peak pairs with a 18 Da  
mass difference.  
3. To remove all pairs results, click Clear List.  
4. To remove a mass from the Spec Peak List (to simplify the  
sequence interpretation), select an entry in the pairs  
results, then click Left or Right under Remove Mass.  
5. Click Copy to Output Window Result Tab to copy  
results to the Result tab. You can then copy from the  
Result tab to another application such as Notepad, or  
print, as needed.  
Sequence To list ion sequences:  
C
1. Click the Sequence tab.  
Figure 3-12 Peptide Fragmentation Sequence  
Data ExplorerSoftware Users Guide  
C-11  
   
Appendix C Data Explorer Toolbox (Visual Basic Macros)  
2. If desired, click Label Immonium Ions.  
NOTE: Label immonium ions before selecting a  
reference peak and starting the search. If you click  
Label Immonium Ions after selecting a reference peak,  
the amino acid labels applied to the spectrum are  
erased, and mass labels are reapplied.  
3. Select a peak in the Spec Peak List, then click Set  
Selected Peak as Reference.  
The software evaluates mass differences between the  
reference peak (labeled with *) and all other peaks, and  
labels any differences that correspond to amino acid  
residues.  
4. Select a result peak from the results list, if desired, then  
continue the sequence search by clicking the Search  
Down or Up buttons.  
C
The newly selected peak is now labeled as the reference  
peak, and mass differences minus (if you click Down) or  
plus (if you click Up) are labeled if they correspond to an  
amino acid residue. You can generate the complete  
sequence ladder by appending new results to existing  
results.  
NOTE: When you search using reference peaks  
selected from the Sequence Results list, new results  
retain the label of the original reference peak, and new  
labels are appended. For example, if you select  
433.08(+Thr)(*) as the reference, results are listed as  
362.13(+Thr)(-Ala), 331.98(+Thr)(-Thr), and  
334.07(+Thr)(-Val). The (+Thr) immediately following  
the mass corresponds to the selected reference (the  
result of the original search). The residue following  
(+Thr) is the residue that corresponds to the mass  
difference between the mass and the reference mass.  
C-12 Applied Biosystems  
Using the Peptide Fragmentation Toolbox  
5. Click Copy to Output Window Result Tab to copy  
results to the Result tab. You can then copy from the  
Result tab to another application such as Notepad, or  
print, as needed.  
Correlation To list ion pair correlations:  
1. Click the Correlation tab.  
2. Select a peak in the Spec Peak List.  
3. Select the correlation you want to determine for the  
selected peak (that is, click a and b if you want to  
determine if the selected peak is a member of an ab  
fragment ion pair).  
C
Figure 3-13 Peptide Fragmentation Correlation  
Data ExplorerSoftware Users Guide  
C-13  
Appendix C Data Explorer Toolbox (Visual Basic Macros)  
4. Click Find Correlation.  
Correlations for the selected peak are listed.  
NOTE: Yp, Sp and Tp represent phosphotyrosine,  
phosphoserine, and phosphothreonine, respectively.  
5. Click Copy to Output Window Result Tab to copy  
results to the Result tab. You can then copy from the  
Result tab to another application such as Notepad, as  
needed.  
6. You can remove a peak from the Spec Peak list by  
selecting a correlation in the results list, then clicking  
Delete Selected Peaks from Spec List.  
Displaying the To display the original labels, select Peak Label from the  
Peaks menu, then deselect User Labels.  
original labels  
C
C-14 Applied Biosystems  
Using the Polymer Analysis Toolbox  
C.6 Using the  
Polymer Analysis Toolbox  
Use the Polymer Analysis Toolbox to determine the following  
values that define the molecular distribution of a polymer:  
M
M
nNumber average molecular weight  
wWeight average molecular weight  
Mzz-average molecular weight  
Mw/MnPolydispersity Index  
The Polydispersity Index represents how widely dispersed the  
polymeric distribution is. A lower value (for example, 1.02)  
indicates a narrowly dispersed polymer. A higher value (for  
example, 3.0) indicates a widely dispersed polymer.  
Using Polymer 1. In the Toolbox Palette dialog box, click Polymer  
Analysis Toolbox. The range displayed in the  
Spectrum window is reflected in the X Display Range  
fields.  
Analysis  
C
NOTE: If you zoom on a different region in the  
Spectrum window, the new range is not updated in the  
X Display Range field, but is used for the analysis when  
you click Calculate. The range is updated when you  
click Calculate.  
2. Baseline correct the trace.  
Peak intensities and areas used in the polymer  
calculations are calculated from a 0 y-axis value, not  
from the calculated baseline of the peak.  
Data ExplorerSoftware Users Guide  
C-15  
     
Appendix C Data Explorer Toolbox (Visual Basic Macros)  
3. Select the mode for the analysis:  
Use the entire mass rangeCalculates  
average molecular weights using all peak  
intensities within the X Display Range. It does  
not distinguish between different polymeric  
species that may be present in the mass  
range.  
Enter an adduct mass.  
C
Figure 3-14 Polymer Analysis Toolbox  
NOTE: If the predominant ion series is M+Na+, enter 23 for  
the adduct ion. If the predominant ion series is M+Ag+,  
enter 108 for the adduct ion. If you are analyzing an  
unknown or a mixture, leave this field blank; the calculation  
is performed without subtracting the adduct ion mass.  
C-16 Applied Biosystems  
Using the Polymer Analysis Toolbox  
Use labeled peaksCalculates average  
molecular weights using areas of peaks listed  
in the Spec Peak List. It allows calculation of  
values for a distinct polymer series when two  
or more species are present.  
Enter an adduct mass.  
Enter the mass of the end group to subtract  
from the calculation. You can leave this field  
blank if you do not know the end group mass.  
CAUTION: Correct labeling of the peaks is essential when  
using the labeled peaks option.  
4. Click Calculate.  
5. Click Copy to Output Window Result Tab to copy  
results to the Result tab. You can then copy from the  
Result tab to another application such as Notepad, as  
needed.  
C
Calculations  
Data ExplorerSoftware Users Guide  
C-17  
Appendix C Data Explorer Toolbox (Visual Basic Macros)  
C.7 Using MS Fit/MS Tag Toolbox  
Use the MS Fit/MS Tag toolbox when analyzing protein digest,  
peptide, or peptide fragment spectra to perform a protein  
database search.  
Preparing data Before using the MS Fit/MS Tag toolbox:  
before accessing  
1. In the Data Explorer software, smooth the data, perform  
a baseline correction, then deisotope.  
2. If needed, use Advanced Peak Detection parameters to  
adjust peak detection ranges and thresholds to screen out  
noise in low-mass regions and to detect peaks in  
high-mass regions.  
3. Set the Set the %Base Peak Intensity or %Max Peak Area  
in peak detection until the unwanted noise peaks are  
removed from the Spec Peak list.  
Running To run the MS Fit/MS Tag macro:  
C
MS Fit/MS tag  
1. In the Toolbox Palette dialog box, click MS Fit/MS Tag.  
Figure 3-15 MS Fit Dialog BoxMS Tag Dialog Box  
C-18 Applied Biosystems  
   
Using MS Fit/MS Tag Toolbox  
2. Click:  
MS-Fit tab If you are examining peptide data  
from a protein digest.  
MS-Tag tab If you are examining PSD data.  
3. Navigate to the web site containing the database to  
search.  
4. Adjust settings on the web site as needed.  
5. Minimize the web page. Do not exit.  
6. In the MS Tag dialog box, enter Parent Ion mass if you  
are using MS-Tag.  
7. Click Copy Peaks.  
The Spec Peak list for the data file is copied to the web  
page.  
8. Start the search from the web site.  
C
The database search is performed. Results of the  
search are displayed on the web site.  
Data ExplorerSoftware Users Guide  
C-19  
Appendix C Data Explorer Toolbox (Visual Basic Macros)  
C
C-20 Applied Biosystems  
Index  
E
Accelerating Voltage  
changes compensated for by  
system 5-25  
Numerics and Symbols  
- in spectrum header 2-31  
%Base Peak Intensity  
definition, chromatogram 3-20  
definition, spectrum 3-22  
in peak detection algorithm 3-67  
Mariner data 3-7  
effect on calibration 5-25  
Accumulate spectra 4-22  
Acquiring data  
comment, displaying 1-15  
comment, displaying for open data  
file 1-15  
comment, displaying when opening  
a data file 2-3  
setting for active detection range,  
spectrum 3-30  
setting global, spectrum 3-22  
setting with data cursor,  
chromatogram 3-11, 3-16  
setting with data cursor,  
spectrum 3-23  
version of software used for 1-15  
Acquisition comment  
displaying for open data file 1-15  
displaying when opening a data  
file 2-3  
setting, chromatogram 3-20  
Voyager data 3-10  
Activate file 2-8  
%Max Peak Area  
Add mode, setting for added  
traces 2-18  
definition, chromatogram 3-20  
definition, spectrum 3-23  
Mariner data 3-7  
setting for active detection range,  
spectrum 3-30  
Adding and subtracting spectra  
from different data files 5-64  
from same data file 5-4  
setting global, spectrum 3-23  
setting, chromatogram 3-20  
Voyager data 3-9  
raw data only 4-20, 5-4  
raw or processed data 5-64  
within a data file 4-20  
+ in spectrum header 2-31  
/ in spectrum header 2-31  
Adding text to traces 2-29  
Adding traces  
maximum number 2-18  
procedure 2-18  
Replace mode, setting 2-18  
A
a and b ion pairs, labeling C-10  
a, b, y -17 ion pairs, labeling C-10  
Advanced baseline correction 5-48  
Advanced peak detection parameters  
description, spectrum 3-28  
setting spectrum 3-17  
Absolute Threshold, replaced by  
Minimum Intensity 3-30  
AC in spectrum header 2-31, 5-34  
AdvBC in spectrum header 2-31, 5-53  
Air Temperature, displaying trace 4-2  
Data Explorer Software Users Guide  
Index-1  
Amino acids, labeling C-5  
Analog signal, displaying 4-2  
Analyzer Temperature, displaying  
trace 4-2  
ANGIOTENSIN_FRAGMENTS.REF 5-  
18  
Automatic Calibration  
On command dimmed 5-34  
see also Calibrating mass,  
automatic (Mariner data  
only)  
I
N
D
E
X
settings (reference masses) for  
Mariner Sequence Control  
Panel 5-27  
settings (reference masses) for  
Voyager Sequence Control  
Panel 5-28  
Annotating traces  
adding text 2-29  
deleting text 2-29  
text from previous trace  
displayed 2-29  
with ASCII text 2-28  
with results 2-28  
settings, importing 5-29  
settings, specifying 5-29  
Automatic macros 6-45  
Apex mass  
copying from peak list 1-41  
labeling 3-57  
AutoSaturation Correction (Mariner  
data only)  
Applied Biosystems  
description 5-62  
Technical Support 9-2  
world wide web address xiv  
effect on Mariner .RST files 5-62  
effect on mass accuracy 5-62  
Area, peak  
Average mass  
detection threshold,  
chromatogram 3-20  
detection threshold, global,  
spectrum 3-23  
detection threshold, local,  
spectrum 3-30  
definition B-6  
labeling partially resolved  
isotopes 3-10, 7-11  
Averaging data, graphic  
compression 1-28  
Averaging spectra 4-22  
displaying 3-38  
Axes  
in peak detection algorithm 3-68  
labeling, chromatogram 3-55  
labeling, spectrum 3-58  
numbers do not match spectrum  
numbers 4-25  
offsetting Y 4-27  
scaling 2-11  
turning off right axis 2-12  
ASC 5-63  
ASC in spectrum header 2-6, 2-31,  
5-63  
ASCII text  
annotating traces with 2-28  
exporting traces to 1-34  
importing traces from 1-35  
B
Background color  
changing 1-23  
Assign Macro 6-38  
default, Mariner 1-4  
default, Voyager 1-4  
Auto Cal, see Calibrating mass,  
automatic (Mariner data only)  
Background signal  
evaluating 7-8  
Autocolor  
description 2-27  
subtracting 4-20  
setting 1-26  
Index-2  
Applied Biosystems  
Bar mode, traces 1-28  
BP Relative% label on data cursor 1-27  
I
BAS files for macros 6-43  
see also Base peak intensity  
in chromatogram header 2-30  
N
Base mass, labeling peaks 3-55, 3-56  
Base peak intensity  
definition 4-2  
scaling to 2-12  
threshold for peak detection,  
chromatogram 3-20  
threshold for peak detection,  
spectrum 3-22, 3-30  
CAL files  
see also Calibration constants  
description 1-7  
exporting during calibration 5-16  
exporting from DAT 1-36  
importing 5-16, 8-20  
importing, error displayed 9-11,  
9-12  
Baseline  
changing line width 1-26  
displaying on trace 3-55, 3-58  
Baseline correction  
chromatogram 4-29  
spectrum 5-47  
spectrum, advanced 5-48  
PSD, creating 8-20  
PSD, overview of creating 8-10  
saving 5-16  
Baseline offset  
chromatogram 4-27  
spectrum 5-45  
Calculator tools  
Basic peak detection parameters  
description, chromatogram 3-19  
description, spectrum 3-22  
resetting 3-18  
elemental composition 6-2  
isotope 6-13  
mass resolution 6-20  
signal-to-noise ratio 6-23  
setting, chromatogram 3-11  
setting, spectrum 3-13  
Calibrating mass, automatic (Mariner  
data only)  
software 1-1, 1-2  
see also Calibration constants  
applying settings to other data  
files 5-29  
calibration reference file  
(REF) 5-17  
calibration settings 5-29  
commands dimmed on menu 9-11  
constants, calculating A and B 5-34  
error, fit 5-12  
BC  
in chromatogram header 2-30, 4-29  
in spectrum header 2-31, 5-47  
BIC files  
see also Instrument settings  
description 1-6  
exporting from DAT 1-36  
exporting from RSD and RCD 1-36  
Binning data, graphic  
compression 1-28  
error, initial 5-12  
overview 5-26  
BO  
peak weighting factor 5-33  
results 5-35  
in chromatogram header 2-30  
in spectrum header 2-31, 5-46  
Data Explorer Software Users Guide  
Index-3  
Calibrating mass, automatic (Mariner  
data only) (continued)  
calibration 5-22  
Calibrating mass, PSD (Voyager data  
only)  
I
all segment masses  
N
D
E
X
considered 8-15  
calibration reference file  
(REF) 8-21  
constants, applying A and B to new  
file 8-20  
error message when selecting  
calibration reference  
file 8-14  
troubleshooting 9-9  
turning on 5-34  
when to use 5-28  
Calibrating mass, manual  
see also Calibration constants  
baseline correcting and deisotoping  
to optimize mass  
internal calibration using known  
monoamino acid fragment  
ions 8-10  
accuracy 7-14  
calibration reference file  
(REF) 5-17  
internal calibration, values  
used 8-7  
multi-point calibration 8-10  
peak detection, checking before  
calibrating 8-10  
commands dimmed on menu 9-11  
constants, applying A and B to new  
file 5-16  
error, fit 5-12  
error, initial 5-12  
peak matching, automatic 8-15  
peak matching, manual 8-16  
peak weighting factor 8-14  
Precursor mass, changing 8-23  
procedure 8-12  
results 8-18  
reverting to instrument  
fit outliers, eliminating 5-12  
importing, Voyager  
requirements 5-25  
multi-point calibration 5-8  
overview 5-5  
peak matches, sorting 5-11  
peak matching, automatic 5-6  
peak matching, manual 5-6  
peak weighting factor 5-10  
procedure 5-8  
selecting peaks for optimum mass  
accuracy 8-19  
selecting peaks near Max Stitch  
Mass 8-19  
single-point calibration 8-10  
results 5-13  
reverting to instrument  
calibration 5-22  
single-point calibration 5-8  
troubleshooting 9-9  
Calibration, see Calibrating mass  
when to use 5-6  
Index-4  
Applied Biosystems  
Calibration constants  
see also Calibrating mass  
applying to new file 5-16, 8-20  
calculating 5-34  
Centering a peak 2-15  
Centroid mass  
calculated during peak  
detection 3-69  
calculating 3-39  
copying from peak list 1-41  
definition B-6  
determining 5-36  
improving accuracy of calculation  
labeling 3-57  
N
displayed in Output window 5-14,  
8-18  
exporting 1-36  
extracting from DAT file 1-36  
importing from another  
source 5-16, 8-20  
PSD 8-20  
reverting to original  
instrument 5-22  
Centroid Percent, spectrum,  
setting 3-26  
Calibration curve, peak weighting  
factors 5-10, 8-14  
Centroiding  
centroid data file size smaller than  
profile 1-33  
Calibration fit error 5-12  
Calibration reference file (REF)  
F 5-18  
command not displayed 9-8  
converting profile to centroid  
data 1-33  
contents 5-18  
creating 5-18  
default 5-18  
duplicate m/z entries 5-20, 5-21,  
9-10  
creating centroid trace 5-36  
during peak detection, improving  
calculation 3-26  
example 5-36  
error message when selecting 8-14  
MARINER_NEG.REF 5-18  
MARINER_POS.REF 5-18  
modifying 5-20  
narrow peaks, specifying as  
Resolved Isotope  
mass 5-21, 9-9  
not listed 9-10  
provided 5-18  
PSD, creating 8-21  
PSD, specifying Resolved Isotope  
masses 8-13  
selecting for automatic  
calibration 5-30  
selecting for manual calibration 5-9  
selecting for PSD calibration 8-13  
viewing contents of 5-10  
VOYAGER.REF 5-18  
histogram trace 5-36  
manual 5-36  
CGM files 1-31, 1-33  
Channel, displaying DAD data by 1-12  
Charge state, peak  
and isotope spacing 3-53  
converting to singly charged 5-59  
converting to zero charged 5-40  
custom label applied only if  
specified charge is  
present 3-62  
determining ion polarity 3-58  
examples 3-32  
filtering peak list by 3-42  
labeled incorrectly 3-60, 9-20  
labels 3-53, 3-58, 3-62  
labels not displayed 3-60, 9-19  
parameters, setting 3-27  
Data Explorer Software Users Guide  
Index-5  
Charge state, peak (continued)  
requirements for labeling 3-53  
single 5-59  
Chromatogram window (continued)  
multispectrum Voyager data,  
displaying 1-13  
I
N
D
E
X
troubleshooting 9-17  
z labels 3-58  
noise filtering 4-17  
peak labels 3-52  
peak list 3-38  
zero 3-40, 3-43  
results, opening 2-40  
retention time, displaying 1-12  
smoothing 4-17  
spectrum number, displaying 1-12  
traces, selecting type 4-2  
vial #, displaying 3-55  
Voyager data, displaying  
multispectrum 2-7  
Chro peak list tab, Output window 1-15  
CHRO window, see Chromatogram  
in peak detection algorithm 3-68  
Chromatogram window  
see also Chromatogram window,  
traces  
Chromatogram window, traces  
adding 2-18  
do not print 2-34  
overlaying 2-25  
overlaying from different data  
previewing and printing 2-33  
trace labels in chromatogram 2-30  
see also Peak Labels  
see also Traces  
adding and subtracting  
spectra 4-20  
baseline correction 4-29  
baseline offset 4-27  
composite PSD data,  
displaying 1-13  
CID data, labeling C-9  
see also Extracted ion  
chromatogram (CNL)  
in chromatogram header 2-30  
constant neutral loss (CNL)  
chromatogram, creating  
from 4-9  
DAD (diode array detector),  
displaying Mariner data 4-2  
description 1-12  
display range, adjusting 2-11  
event tags (Mariner data only),  
filtering 4-24  
extracted absorbance  
chromatogram (XAC) from  
Mariner data, creating  
from 4-13  
extracted ion chromatogram (XIC)  
chromatogram, creating  
from 4-5  
Colors  
background, changing 1-23  
printing 2-33  
customizing 1-25  
default, Mariner 1-4  
default, Voyager 1-4  
overlaid traces 2-27  
setting 1-25  
Combined spectrum  
creating 5-4  
definition 5-2  
example 7-4, 7-6  
label 5-4  
improving signal-to-noise ratio 7-2  
Mariner data, displaying 1-12, 2-2  
Index-6  
Applied Biosystems  
CombiSolv data  
all peaks 1-40  
displaying one injection 4-24  
Event Tag Filtering command  
dimmed (Mariner data  
only) 4-24  
apex mass list 1-41  
centroid mass list 1-41  
data from other data files 2-37  
data to Windows clipboard 1-38  
displayed peaks 1-39  
mass list 1-41  
peak list 1-40  
results 2-28  
to metafile 1-38  
trace data 1-39, 2-37  
Comment  
acquisition, displaying 1-15  
acquisition, displaying for open  
data file 1-15  
acquisition, displaying when  
opening a data file 2-3  
result file 1-15, 2-40  
Comparing data files 2-38  
Correcting baseline  
chromatogram 4-29  
spectrum 5-47  
Composite spectrum, PSD  
automatically processed when  
generated 8-6  
spectrum, advanced 5-48  
displaying segment traces 8-3  
generating 8-2, 8-5  
how it is generated 8-6  
labeling fragment ions 8-8  
Max Stitch Masses 8-4  
precursor mass, changing 8-23  
Correlation factor, noise filtering 4-18,  
5-43  
CT  
file suffix 1-33  
in spectrum header 2-32, 5-36  
Composition, elemental 6-2  
Compression, data 1-28  
CTS files 1-7  
disabling 2-23  
purging 2-23  
Computer  
Cursors, see Data cursors 1-27  
technical support for altered  
configuration A-1  
Custom labels  
see Graphic options, peak labels,  
customizing  
see Peak labels, custom  
Configuration, exporting from DAT  
file 1-36  
Customer feedback, Technical  
Publications e-mail address xiv  
Constant neutral loss (CNL), see  
Extracted ion chromatogram  
(CNL)  
colors, text, and line width 1-25  
creating your own software  
applications 1-5  
data compression 1-28  
Data Explorer window 1-17  
defaults 1-17  
Context-sensitive menus 1-14  
Converting  
see also Exporting  
DAT to ASCII 1-34  
profile data to centroid 1-33  
SPC to DAT 1-30  
graphic options 1-24  
peak labels 3-52  
Voyager .MS files 1-30  
processing settings 1-17  
Data Explorer Software Users Guide  
Index-7  
Customizing (continued)  
SET files 1-17  
Data cursors (continued)  
turning on and off in graphic  
options 1-27  
I
toolbars 1-22  
N
D
E
X
turning on and off in peak  
detection 3-23  
vertical 1-27  
D
Data Explorer  
DAD  
moving between open files 2-8  
starting and exiting the  
software 1-3  
window, customizing 1-17  
window, parts of 1-11  
displaying Channels 1-12, 4-2  
displaying TAC 1-12  
displaying traces 2-6  
extracted absorbance  
chromatogram (XAC) 4-13  
in spectrum header 2-32  
spectrum, displaying 4-2, 5-2  
TAC in chromatogram header 2-30  
XAC in chromatogram header 2-30  
Data Explorer examples  
Mariner data 7-2  
Voyager data 7-11  
Data Explorer Toolbox  
Dark background  
changing colors 1-20  
default settings 1-23  
SET file 1-19  
Ladder Sequencing macro C-2, C-5  
modifying C-2  
MS Fit/MS Tag macro C-2  
overview C-2  
Peptide fragmentation macro C-2,  
C-9  
Polymer analysis macro C-2, C-15  
preparing data for C-3  
Visual Basic References  
required C-3  
DAT format  
converting DAT file to ASCII 1-34  
converting from SPC 1-30  
converting profile to centroid 1-33  
extracting information from 1-6,  
1-36  
file properties, adding to DAT  
file 1-31  
file properties, searching 1-32  
file properties, viewing 1-32  
information stored in 1-6  
overview 1-5  
Data file  
activate 2-8  
cannot find 9-3  
closing 2-10  
closing and running a macro  
automatically 6-45  
comparing open files 2-36, 2-38  
converting from profile to  
centroid 1-33  
converting to ASCII 1-34  
copying traces from 2-37  
decreasing size 1-33  
default settings, applying 2-4  
error message displayed when  
opening PSD 8-2, 9-3  
results, saving, opening,  
deleting 2-38  
Data cursors  
%Base Peak Intensity,  
spectrum 3-23  
displaying retention time on 1-27  
horizontal 1-27  
labels 1-27, 4-2  
printing and suppressing 1-27  
Index-8  
Applied Biosystems  
Data file (continued)  
full name not displayed 1-14  
inserting traces into 2-37  
moving between open 2-8  
multiple, zooming 2-36  
name 1-14  
deconvolution  
Default settings, applying from SET  
file 2-4  
E
DEFAULTBLACK.SET 1-19, 1-23  
Defaults  
names do not print 9-13  
opening 1-30  
opening and applying default and  
selected settings 2-4  
opening and running a macro  
automatically 6-45  
opening manually 2-2  
opening PSD 8-2  
read-only 2-7  
result, opening 2-39, 2-40  
trace browser 2-8  
version of software used to  
acquire 1-15  
background color 1-23  
colors, Mariner 1-4  
colors, Voyager 1-4  
peak detection 3-2  
processing and graphics settings,  
data file 2-4  
REF files 5-6, 5-18  
SET files 1-19, 1-23, 8-11  
trace color 1-23  
DEFAULTWHITE.SET 1-19, 1-23  
description 3-45  
enhancing peak interpretation 3-9  
formula 3-49  
procedure 3-48  
working with separate 2-36  
zooming multiple 2-36  
Data format, see DAT format  
Data points  
copying 1-39  
requirements 3-47  
determining number across a  
eliminating from spectrum 5-56  
Deleting  
peak labels from the trace 3-44,  
3-59  
peaks from the peak list 3-44, 3-59  
text annotation 2-29  
Database search macro C-18  
DATAEXPLORER.VB6  
exporting macros from 6-44  
importing new macros into 6-43  
location 6-35  
Detection Ranges  
adding, chromatogram 3-19  
adding, spectrum 3-29  
calculating automatically 3-4  
combining, chromatogram 3-20  
combining, spectrum 3-29  
deleting chromatogram 3-20  
deleting, spectrum 3-29  
example 3-4  
new supplied macros not available  
until imported 6-43  
not overwritten when new software  
DBE, definition 6-6  
Decimal points, number  
displayed 3-55, 3-56  
overlapping regions 3-5  
PSD data 8-11  
DECONV in spectrum header 2-32,  
5-40, 5-41  
Data Explorer Software Users Guide  
Index-9  
Detection Ranges (continued)  
Elemental composition (continued)  
Isotope Match Score 6-6  
Isotope Match Score not reported  
I
setting manually,  
N
D
E
X
setting manually, spectrum 3-28  
setting parameters globally,  
spectrum 3-30  
isotope min and max values  
ignored 6-8, 6-11  
limits, setting 6-7, 6-9, 6-12  
Periodic table 6-10  
procedure 6-3  
Detection, see Peak detection  
DI in spectrum header 2-32, 3-49  
Diode array detector data, see DAD  
results 6-5  
Elemental targeting  
description 6-31  
Diode array detector data, see DAD  
data  
Isotope Match Score 6-33  
procedure 6-31  
results 6-33  
Disk space, conserving by converting  
profile data to centroid 1-33  
Display range  
theoretical isotope, displaying 6-33  
scaling 2-11  
x range, expanding 2-11  
y range, expanding 2-11  
Eliminate Fit Outlier 5-12  
E-mail address, Technical  
Publications xiv  
Display Trace dialog box 2-18  
Displaying  
PSD calibration 8-6  
resolution-based peak  
detection 3-4  
peak labels 3-54, 3-56  
Voyager chromatogram window 2-7  
DNA residues, labeling C-5  
Error, calibration  
fit 5-12  
Double bond equivalents (DBE)  
definition 6-6  
initial 5-12  
determining 6-6  
Event tag filtering (Mariner data only)  
not displayed on Process  
menu 4-24  
Duplicating a trace 2-15  
performing 4-24  
redisplaying original data 4-26  
E
EF  
Event tag, MS Method (Mariner data  
only)  
in chromatogram header 2-30, 4-25  
redisplaying original data 4-26  
only), events  
determining tags in file 4-23  
displaying 4-23  
filtering 4-23  
redisplaying original data 4-26  
Elemental composition  
calculating 6-2  
description 6-2  
elements, adding 6-9  
fragment ions, electron state for  
accurate results 6-5  
Examples, see Data Explorer examples  
fragment ions, identifying 6-2  
Index-10 Applied Biosystems  
Excel, see Microsoft Excel  
Exiting software 1-3  
I
Failed to create empty document  
message 9-3  
Expanding traces 2-21  
Exporting  
Failed to open chromatogram data  
message 9-4  
see also Converting  
ASCII data 1-34  
BIC 1-36  
File format, converting SPC to  
DAT 1-30  
CAL files 1-36  
Configuration from DAT file 1-36  
entire data file 1-34  
macros from DATA  
EXPLORER.VB6 6-44  
MSM and CAL from DAT 1-36  
results 2-39  
File management 1-30  
File name  
does not print 9-13  
full name not displayed 1-14  
File properties  
searching 1-32  
viewing 1-32  
RSD and RCD files 2-39  
trace to ASCII format 1-34  
File size, reducing 1-33  
File types 1-6 to 1-8  
Extracted absorbance chromatogram  
(XAC) (Mariner data only)  
creating 4-13  
Files  
automatically running macros when  
opening or closing 6-45  
cannot find 9-3  
definition 4-3  
displaying 4-13  
trace label 4-14, 4-15  
closing 2-10  
Extracted ion chromatogram (CNL)  
corresponds to loss of  
fragments 4-9  
error message displayed when  
opening PSD 8-2, 9-3  
opening manually 2-2  
opening PSD 8-2  
creating, example 4-12  
definition 4-3, 4-9  
determining if mass difference 4-9  
trace label 4-9, 4-10  
read-only 2-7  
Filter Width  
Extracted ion chromatogram (XIC)  
creating 4-5  
in peak detection algorithm 3-3,  
3-67  
setting automatically 3-3  
setting manually,  
chromatogram 3-20, 3-21  
too high, peaks not labeled 3-20,  
3-21, 3-31, 9-19  
definition 4-3, 4-4  
example 7-2, 7-5, 7-10  
improving signal-to-noise ratio 4-6  
including only mass of interest 4-6  
trace label 2-31, 4-7, 4-8, 4-16  
Extraction mode, setting 4-7, 4-11  
value used when resolution-based  
detection enabled 3-3  
Data Explorer Software Users Guide Index-11  
Filter Width Increment  
I
setting manually, spectrum 3-31  
suggested value, spectrum 3-31  
value used when resolution-based  
peak detection  
N
D
E
X
Gaussian Fitting, peak detection 3-26  
Gaussian smoothing, see Smoothing  
Global peak detection parameters  
description, spectrum 3-22  
overriding for individual detection  
ranges, spectrum 3-30  
enabled 3-31  
Filtered traces, viewing, see Filtering  
Filtering  
chromatogram traces 4-23  
event tags (Mariner data only) 4-24  
monoisotopic peaks 3-43  
noise, chromatogram 4-17  
peak list 3-42  
setting spectrum 3-13  
Graphic options  
accessing 1-24  
default settings for white or dark  
background 1-19, 1-23  
extracting from DAT file 1-36  
graphic compression 1-28  
peak labels, customizing 1-25  
saving and restoring 1-20  
saving to a SET file 1-37  
setting graph and plot colors  
with 1-25  
Fit Error, calibration 5-12  
Formulas  
determining if present in observed  
spectrum 6-31  
determining possible for a  
mass 6-2  
Fragment ions  
turning off right axis 2-12  
correspond to loss of 4-9  
determining if mass difference 4-9  
electron state for accurate  
results 6-5  
generating list of masses with Ion  
Fragmentation  
Graphic settings  
applying 2-4  
automatically saved when data file  
closed 1-18  
description 1-18  
extracting from DAT file 1-37  
modifying 1-19  
saving and restoring 1-20  
files 1-19  
identifying with Elemental  
Composition calculator 6-2  
labeling 6-25, C-9  
Peptide fragmentation macro C-9  
Fragment spectrum, PSD, see  
Segment spectrum, PSD  
H
FRM files for macros 6-43  
FWHM 6-20  
Help, see PerSeptive Biosystems  
Technical Support  
Histogram, centroid 5-36  
Horizontal cursor 1-27  
Horizontal peak labels 3-55, 3-58  
How to use this guide xi  
Index-12 Applied Biosystems  
Ion Fragmentation Calculator  
description 6-25  
I
Immonium ions, labeling C-9  
D
PSD segment spectra, labeling 8-8  
REF file, PSD, creating 8-21  
results 6-29  
IMMONIUM_IONS.REF 5-18, 8-19  
Import Calibration  
error displayed 9-11, 9-12  
procedure 5-16  
sequence codes, acceptable 6-26  
ISO in spectrum header 2-32, 6-6,  
6-16, 6-18, 6-33  
PSD 8-20  
Importing  
macros into DATA  
average mass, labeling 3-10, 7-11  
description B-1  
displaying theoretical 6-33  
isotopic envelope B-5  
list of common B-8  
monoisotopic and average  
masses B-6  
monoisotopic, labeling 3-43  
overview B-1  
EXPLORER.VB6 6-43  
PKT files to Excel 3-41  
trace from ASCII format 1-35  
Improving signal-to-noise ratio 5-2, 7-2  
Initial Error, calibration 5-12  
Inserting peaks in peak list 3-39  
In-source CID data, labeling C-9  
Instrument calibration  
reapplying 5-22  
parameters controlling charge state  
determination 3-27  
reverting to 5-22  
partially resolved, labeling 7-11  
resolution, limits of system 3-53  
resolution, overview B-4  
spacing, and charge state 3-53  
theoretical, generating with  
Calculator 6-6  
theoretical, generating with  
Elemental Targeting 6-33  
theoretical, generating with Isotope  
Calculator 6-18  
Instrument settings  
extracting from DAT file 1-36  
tab, Output window 1-16  
viewing 1-16  
Integration  
setting, chromatogram 3-21  
setting, spectrum 3-26  
Valley-to-Baseline,  
Valley-to-Baseline, spectrum 3-26,  
3-70  
troubleshooting 9-17  
amino acid codes 6-14  
charge state to calculate 6-15  
element codes 6-14  
evaluating traces 6-18  
examples 6-16  
chromatogram 3-21, 3-70  
Valley-to-Valley, spectrum 3-26,  
3-70  
Internal mass calibration 5-5, 5-26  
Internal standard calibration, see  
Calibrating mass  
result, failed to calculate 9-6  
results 6-19  
using 6-13  
Data Explorer Software Users Guide Index-13  
Isotope Match Intensity, elemental  
targeting 6-33  
Labels (continued)  
CNL 4-9  
I
CT 5-36  
N
D
E
X
Isotope Match Score  
DECONV 5-40, 5-41  
filtered trace 4-25  
ISO 6-6, 6-16, 6-33  
MC 5-13  
NF 4-19, 5-44  
NR 4-19, 5-44  
RSM 5-44  
SC 5-60  
SM 4-19, 5-44  
TR 5-57  
XAC 4-14, 4-15  
XIC 4-7  
elemental composition 6-6  
elemental composition, not  
reported for fragment ion  
calculations 6-6  
elemental targeting 6-33  
K
Keywords, Windows NT  
entering 1-31  
searching 1-32  
viewing 1-32  
Ladder Sequencing macro C-2, C-5  
lost when you exit 9-13  
L
Labeling peaks  
setting permanently 2-35  
see also Peak labels  
chromatogram 3-54  
customizing label appearance 1-25  
factors affecting 3-52  
manually 3-39  
monoisotopic 3-43  
partially resolved 7-11  
spectrum 3-56  
LBC or LBS files, in DAT format  
applying 3-65  
creating 3-64  
description 1-7  
extracting from DAT file 1-36  
Line mode, traces 1-28  
Line width, setting 1-26  
Linking  
with amino acid, DNA, or RNA  
multiple data files for zooming 2-13  
traces within a data file 2-21  
views 2-13  
with area 3-58  
with time 3-55  
views, does not work 9-7  
with vial number 3-55  
setting, spectrum 3-28  
Labels  
see also Data cursor  
see also Peak labels  
see also Trace labels  
AC 5-34  
added and subtracted spectra 4-22  
AdvBC 5-53  
Loss of H2O, determining C-11  
Loss of NH3, determining C-10  
Low Mass Gate spike, eliminating 5-56,  
5-57  
BC 4-29, 5-47  
BO 4-28, 5-46  
Index-14 Applied Biosystems  
Mariner data (continued)  
examples 7-2  
M
I
in-source CID, labeling C-9  
isotope resolution limits 3-53  
mass accuracy affected by peak  
shape 5-7, 5-47  
Macro Recorder  
N
E
advanced editing 6-42  
buttons, assigning to macros 6-38  
DATAEXPLORER.VB6  
location 6-35  
DATAEXPLORER.VB6 not  
overwritten when new  
software installed 6-43  
deleting a macro 6-41  
description 6-34  
peak detection strategy 3-6  
Mariner Sequence Control Panel,  
automatic calibration settings  
(reference masses) for 5-27  
MARINER.REF  
viewing masses in 5-10  
exporting macros from DATA  
EXPLORER.VB6 6-44  
functions not supported 6-35  
importing macros into DATA  
EXPLORER.VB6 6-43  
location of macros 6-35  
maximum number of macros 6-34  
new supplied macros not available  
until imported 6-43  
MARINER_NEG.REF  
provided 5-18  
provided 5-18  
Mass  
apex, copying from peak list 1-41  
apex, labeling 3-57  
average B-6  
in chromatogram header 2-30  
in chromatogram header, replaced  
by XIC 2-31  
labels 3-56  
peak  
running a macro 6-39  
toolbar, displaying 6-35  
Visual Basic Editor 6-42  
Macros  
provided, see Data Explorer  
Toolbox  
running automatically when  
opening and closing data  
files 6-45  
Mass accuracy  
affected by peak shape 5-7, 5-47  
multi-point calibration versus  
single-point calibration 5-8,  
8-10  
running manually 6-39  
Mass accuracy, optimizing  
(Mariner data only) 5-62  
by baseline correcting and  
deisotoping before  
Manual calibration, see Calibrating  
mass, manual  
Manually labeling peaks 3-39, 3-52  
Mariner data  
Chromatogram window,  
displaying 2-2  
DAD data (diode array detector),  
displaying 1-12  
calibrating 7-14  
by baseline correcting and noise  
filtering before  
calibrating 5-7  
DAD spectrum, displaying 5-2  
Data Explorer Software Users Guide Index-15  
Mass Calibration commands  
dimmed on menu 9-11  
Maximize trace window 2-9  
I
Maximum Charge State  
not displayed on menu 9-8  
N
D
E
X
Mass calibration, see Calibrating mass  
set too low 3-33  
Mass, centroid  
setting 3-27  
calculating 3-39  
copying from peak list 1-41  
definition B-6  
displaying peaks as  
histograms 5-36  
labeling 3-57  
setting too low 3-35  
MC in spectrum header 2-6, 2-32, 5-13,  
5-15, 5-17, 5-22, 5-23, 7-17,  
8-18  
Metafile, copying to 1-39  
Mass deconvolution  
command dimmed on Process  
menu 5-37, 9-8  
Microsoft applications  
Excel 1-5  
Microsoft Office 1-5  
Visual Basic Editor 6-42  
Word 1-5  
example 7-4  
performing 5-37  
result, failed to calculate 9-8  
results 5-40  
Microsoft Excel, importing PKT files  
to 3-41  
Mass Difference peak label  
from adjacent peak (regular  
labels) 3-57  
Minimum Area, setting spectrum 3-30  
Minimum Intensity  
from adjacent peak (user  
labels) 3-62  
setting spectrum 3-30  
from selected peak 3-57  
Mn Mw Mz Mw/Mn, determining C-15  
clipboard 1-41  
deisotoping a spectrum 3-9, 3-45  
displaying in peak list 3-43  
filtering 3-43  
Mass Offset, replaced by Mass  
Difference from Selected  
Peak 3-57  
labeling 3-43  
mass, definition B-6  
reducing trace to 3-9, 3-45  
Mass Resolution Calculator, see  
Resolution, mass  
Mass resolution defaults used in peak  
detection 3-24  
Moving between open files 1-12  
MS Fit/MS Tag macro C-2  
Match Charge State 3-62  
MS Method (Mariner data only)  
applications 4-23  
Max Stitch Mass  
definition 8-4  
displaying 8-4  
conditions and tags,  
displaying 4-23  
optimum focus and resolution  
observed near this  
mass 8-4  
updated when Precursor mass  
changed 8-24  
Event Tag Filtering command  
dimmed 4-24  
event tag, filtering 4-23  
instrument settings, extracting from  
DAT file 1-36  
Index-16 Applied Biosystems  
MS Method (Mariner data only)  
(continued)  
Noise Threshold  
chromatogram, calculated  
automatically 3-21, 3-68  
in peak detection algorithm 3-68  
setting locally, spectrum 3-30  
instrument settings, viewing 4-23  
spectrum numbers in filtered  
trace 4-25  
N
D
X
MSM files  
Noise, screening out  
noise filtering command 4-17, 5-42  
subtracting spectra and creating  
extracted ion  
extracting from DAT file 1-36  
overview 1-7  
Multiple Charge command  
dimmed on menu 5-37, 9-8  
chromatogram (Mariner  
data) 7-8  
truncate spectrum 5-56  
Multiple data files  
Nozzle Temperature, displaying  
trace 4-2  
comparing 2-36, 2-38  
copying traces into a window 2-37  
printing 2-36  
working with separately 2-36  
zooming 2-13, 2-36  
in chromatogram header 2-30, 4-19  
in spectrum header 2-32, 5-44  
files 2-7  
Number average molecular weight,  
determining C-15  
Multiply charged peaks  
deconvoluting 5-37  
Number of data points across a peak,  
determining 3-21, 3-31, 5-51  
single-charge conversion 5-59  
Multi-point calibration, see Calibrating  
mass  
Offsetting baseline  
chromatogram 4-27  
spectrum 5-45  
N
Negative ion z label 3-58  
Offsetting UV trace (Mariner data only)  
Mariner data 4-30  
Neutral loss chromatogram, see  
Extracted ion chromatogram  
(CNL)  
restoring original trace 4-31  
Open files, moving between 2-8  
NF  
Opening  
in chromatogram header 2-30, 4-19  
in spectrum header 2-32, 5-44  
data files 2-2  
data files and automatically running  
a macro 6-45  
Noise filtering  
PSD data files 8-2  
result files 2-40  
chromatogram 4-17  
spectrum 5-42  
spectrum, recommended  
setting 4-18, 5-43  
Data Explorer Software Users Guide Index-17  
Output window  
acquisition comment,  
color, changing 1-25  
I
displaying 1-15  
calibration statistics,  
displaying 5-13, 8-18  
Chro Peak list tab 1-15  
clearing 1-16  
N
D
E
X
see also Peak detection  
see also Peak detection, Mariner  
data  
see also Peak detection,  
resolution-based  
data  
accessing 3-11, 3-13  
Advanced Settings tab not  
Advanced Settings, spectrum 3-17,  
3-28  
closing 1-16  
copying results from 2-28  
instrument settings 1-16  
Instrument Settings tab 1-15  
peak list, chro 1-15  
peak list, displaying 1-15  
peak list, importing and saving in  
Excel 3-41  
peak list, saving as a file 3-40  
peak list, spec 1-15  
Results tab 1-15  
results, displaying 1-15  
sample info 1-15  
Sample Info tab 1-15  
Spec Peak list tab 1-15  
baseline, changing line width 1-26  
baseline, displaying 3-55, 3-58  
Basic Settings, spectrum 3-13,  
3-22  
Overlapping peak labels 3-55, 3-58  
charge state determination 3-27  
chromatogram 3-68  
chromatogram, Noise Threshold  
Overlaying traces  
active trace color 2-26  
advancing 2-26  
data cursors, turning on and  
off 3-23  
autocolor 2-27  
colors, setting 2-27  
procedure 2-24  
data points across a peak,  
determining number 3-21,  
3-31, 5-51  
scaling options 2-27  
troubleshooting 9-6  
default 3-2, 3-71  
Filter Width and Increment  
used 3-3  
P
Gaussian Fitting 3-26  
isotope determination 3-27  
isotope, partially resolved 7-11  
manually inserting peaks in peak  
list 3-39  
Peak  
area, see Area, peak  
centroid, see Centroid mass  
charge state, see Charge state,  
peak  
deisotoping, see Deisotoping  
inserting 3-39  
Index-18 Applied Biosystems  
Peak detection (continued)  
Noise Threshold, calculated  
automatically for  
Peak detection, resolution-based  
see also Peak detection  
default resolution 3-6, 3-24  
description 3-3  
D
chromatogram data 3-21,  
3-68  
overview 3-2  
Peak Processing parameters,  
spectrum 3-16, 3-26  
peak start and end,  
displaying 3-55, 3-58  
peaks do not appear in  
spectrum 9-17  
process that occurs during 3-67  
proteins 3-6  
ranges, overlapping 3-5  
regions, setting  
chromatogram 3-19, 3-20  
regions, setting spectrum 3-28  
resetting Basic settings 3-18  
spectrum 3-68  
troubleshooting 9-14  
use same settings for all  
traces 3-21, 3-25  
Filter Width value used 3-3  
formula used to calculate number  
of data points 3-4  
overriding 3-10, 3-14  
Peak detection, Voyager data  
see also Peak detection  
complex digests 7-18  
default settings 3-71  
deisotoping to aid in peak  
detection 3-9  
high mass peaks not detected 3-9  
improving by baseline correcting,  
smoothing 3-8  
noise detected as peaks 3-9  
detected 3-10, 7-11  
Peak detection parameters,  
chromatogram  
strategy for 3-8  
description 3-19  
setting 3-11, 3-19  
Peak integration, see Integration  
Peak detection parameters, spectrum  
Advanced 3-28  
see also Peak labels, charge state  
see also Peak labels, filtering  
see also Peak labels, Mass  
Difference  
45 degree angle 3-55, 3-58  
amino acid C-5  
apex 3-56  
applying user labels from LBC and  
LBS files 3-65  
area, chromatogram 3-55  
area, spectrum 3-56, 3-58  
base peak mass 3-55, 3-56  
baseline, changing line width 1-26  
baseline, displaying 3-55, 3-58  
Basic 3-22  
global, description 3-22  
global, setting 3-13  
for 3-30  
local, description 3-28  
local, setting 3-28  
Peak Processing 3-26  
Peak detection, Mariner data  
see also Peak detection  
default settings 3-71  
noise detected as peaks 3-7  
strategy for 3-6  
troubleshooting 3-6  
Data Explorer Software Users Guide Index-19  
Peak labels (continued)  
centroid 3-56  
Peak labels, filtering  
charge state 3-43  
I
custom 3-61  
custom, creating for fragment  
spectra 8-9  
customizing 1-25  
decimal places displayed 3-55,  
3-56  
deleting from trace 3-44, 3-59  
displaying 3-65  
DNA C-5  
extracting from DAT file 3-64  
factors affecting 3-52  
height 3-55, 3-56  
monoisotopic peak 3-43  
N
D
E
X
Peak labels, Mass Difference  
From Adjacent Peak (regular  
labels) 3-57  
From Adjacent Peak (user  
labels) 3-62  
From Selected Peak 3-57  
Peak list  
charge state, displaying  
selected 3-42  
Chro and Spec 3-40  
contents 3-38  
copying apex masses only 1-41  
copying with headings 1-40  
copying without headings 3-41  
deleting peaks 3-44, 3-59  
description 3-37  
displaying 3-37, 3-40  
filtering 3-43  
importing and saving in Excel 3-41  
inserting peaks manually 3-39  
monoisotopic peaks,  
displaying 3-42  
immonium ions C-9  
manually applying 3-39, 3-52  
manually inserting peaks 3-39  
monoisotopic 3-43  
not displayed 3-59, 3-66, 9-14,  
9-15  
overlapping 3-55, 3-58  
peak start and end,  
displaying 3-55, 3-58  
PSD 8-8  
RNA C-5  
spectrum number 3-55, 3-56  
spectrum, setting 3-56  
time 3-55, 3-56  
troubleshooting 9-14  
turning on and off in  
chromatogram 3-52  
user defined 3-61  
printing 3-44  
saving as a file 3-40  
sorting 3-42  
when created 3-37  
zero charge state displayed 3-43  
vertical 3-55, 3-58  
Vial number 3-55  
sorting 5-11  
Peak labels, charge state  
filtering 3-43  
automatic calibration 5-32  
manual calibration 5-6, 5-9  
PSD calibration 8-14  
incorrect 9-20  
not displayed 9-19  
parameters used 3-53  
requirements 3-53  
selecting 3-58  
Peak resolution, see Resolution, mass  
Peak Threshold%, replaced by %Base  
Peak Intensity 3-20, 3-22  
user labels 3-62  
Index-20 Applied Biosystems  
Peak weighting factors 5-10, 8-14  
Process menu, commands not  
displayed 9-8  
Peak Width  
minimum and maximum used 3-21,  
Processing  
3-25  
see also Processing settings  
menu 9-6  
set automatically by software 3-21,  
3-25  
troubleshooting 9-6  
Peaks, do not appear in spectrum 9-17  
Peptide fragmentation macro C-2, C-9  
Periodic table 6-10  
Processing History  
description 2-22  
disabling 2-23  
options 2-23  
PerSeptive Biosystems Technical  
Support, see Applied  
using 2-22  
Biosystems Technical Support  
Processing settings  
PKT files 3-40  
automatically saved when data file  
closed 1-18  
Points across a peak,  
Polydispersity index, determining C-15  
Polymer analysis macro C-2, C-15  
Positive ion z label 3-58  
customizing 1-19  
description 1-18  
extracting from DAT file 1-36  
modifying 1-19  
peak detection parameters,  
spectrum 3-26  
saving for use with other data  
files 1-19  
Precursor mass, see PSD analysis  
Preface xi  
Previewing traces before printing 2-33  
Print preview 2-33  
Print Setup 2-36  
Printer, setting to landscape  
orientation 2-35  
Profile data  
converting to centroid 1-33  
traces do not print 1-26, 2-35  
Printing  
all traces in view 2-34  
changing colors to black  
before 2-33  
data cursors 1-27  
data file names do not print 2-34,  
2-37  
landscape orientation 2-35  
multiple files 2-34  
peak list 3-44  
Protein database search macro C-18  
Proteins, detection 3-6  
Proton mass, value 3-32  
see also Calibration mass, PSD  
(Voyager data only)  
see also Voyager Biospectrometry  
Workstation Users Guide  
CAL file, creating 8-20  
calibrating 8-10  
traces 2-33  
traces do not print 2-34  
troubleshooting 9-13  
without previewing 2-34  
Change Mass function 8-23  
Data Explorer Software Users Guide Index-21  
PSD analysis (continued)  
Realign, UV trace 4-30  
I
composite spectrum,  
Recording a macro 6-37  
displaying 8-2, 8-5  
composite spectrum, how it is  
generated 8-6  
N
D
E
X
REF files, see Calibration reference file  
(REF)  
fragment labels, applying 8-8  
optimum resolution observed near  
Max Stitch Mass 8-4  
peak detection algorithm 3-5  
peak detection, checking 8-10  
Peptide fragmentation macro C-9  
precursor mass, changing 8-23  
PSD, overview of creating 8-10  
REF file, creating 8-21  
Reference masses, calibration  
see also Calibration reference file  
adding to REF file 5-18  
displaying list of 5-10, 8-16  
Panel 5-27, 5-28  
selecting 5-10, 5-14, 7-17, 8-15,  
8-16  
segment labels, applying 8-8  
segment traces, displaying 8-3  
References required for Data Explorer  
Toolbox (Visual Basic macros  
provided) C-3  
PSD calibration  
equation 8-6  
Precursor mass, impact of  
changing 8-23  
Related documents xiv  
Removing traces  
active 2-21  
PSD data, in Voyager, displaying 1-13  
inactive 2-21  
PSD mode  
Replace mode, setting for added  
traces 2-18  
constants 8-6  
equation 8-6  
Resolution Calculator not displayed 9-8  
Resolution, mass  
R
calculating 6-20  
command not on menu 9-8  
default peak height used 6-20  
defaults used in peak  
detection 3-24  
Raw data, copying x,y pairs 1-39  
RCD file  
deleting 2-39  
description 1-10  
exporting 2-39  
opening 2-39  
isotope B-4  
isotope, Mariner data 3-53  
isotope, Voyager data 3-53  
PSD segment, optimum observed  
near Max Stitch Mass 8-4  
results 6-22  
RCT files  
see also Results  
name of raw data file result is  
derived from 2-38, 2-39,  
2-41  
opening 2-39, 2-40  
saving 2-40  
trend within PSD segments 8-11  
Resolution-based peak detection, see  
Peak detection,  
resolution-based  
Read-only files, viewing 2-7  
Index-22 Applied Biosystems  
Result tab, Output window 1-15  
RSD file  
deleting 2-39  
Results  
exporting 2-39  
extracting information from 1-36  
opening 2-39  
see also RSD and RCD  
annotating traces with 2-28  
copying 2-28  
displaying in Output window 1-15  
elemental composition 6-5  
elemental targeting 6-33  
exporting 2-39  
extracting information from 1-36  
ion fragmentation 6-29  
isotope 6-19  
D
RSM in spectrum header 2-32, 5-44  
RST files  
see also Results  
name of raw data file result is  
derived from 2-38, 2-39,  
2-41  
mass deconvolution 5-40  
name of raw data file result is  
derived from 1-15  
not affected by AutoSaturation  
Correction 5-62  
opening 2-39, 2-40  
saving 2-40  
peak list 3-38  
RCD and RSD files 2-39  
resolution, mass 6-22  
signal-to-noise ratio 6-24  
Running a macro  
automatically 6-45  
manually 6-39  
Results (DAT)  
copying 2-28  
deleting results 2-38  
opening results 2-38  
saving results 2-38  
title when saving 2-38  
Saturated spectra, correcting (Mariner  
data only) 5-62  
Results (SPC)  
SC in spectrum header 2-32, 5-60  
AutoSaturation Correction, effect  
copying 2-28  
name of raw data file result is  
derived from 2-38, 2-39,  
2-41  
Scaling  
see also Display Range  
overlaid traces 2-27  
to Absolute Value 2-12  
to Base Peak 2-12  
to min/max Y 2-12  
opening RST and RCT files 2-39,  
2-40  
Searching for keywords 1-32  
saving RST and RCT files 2-40  
Retention time, displaying  
for chromatogram 1-12  
on data cursor 1-27  
creating custom peak labels 8-9  
labeling 8-8  
lower masses require higher Filter  
Width setting 8-11  
optimum resolution observed near  
Max Stitch Mass 8-4  
resolution trend within 8-11  
Reverting to instrument  
calibration 5-22  
Right axis, turning off 2-12  
RNA residues, labeling C-5  
Data Explorer Software Users Guide Index-23  
Sequence Control Panel, Mariner,  
automatic calibration settings  
(reference masses) for 5-27  
Single-charge conversion  
description 5-59  
I
multiply-charged peaks seen after  
conversion 5-62  
procedure 5-59  
troubleshooting 5-62  
N
D
E
X
Sequence Control Panel, Voyager,  
automatic calibration settings  
(reference masses) for 5-28  
SET files  
Single-point calibration, see Calibrating  
mass  
applying 1-20  
contents 1-18  
creating for Mariner Sequence  
Control Panel 5-27  
creating for Voyager Sequence  
Control Panel 5-28, 5-29  
customizing 1-19  
defaults 1-19, 1-23  
description 1-7  
extracting from DAT 1-37  
modifying 1-19  
restoring 1-20  
Control Panel 5-33  
viewing contents of 1-18  
in spectrum header 2-32, 5-44  
chromatogram 4-17  
setting method performed by  
toolbar 5-42  
spectrum 5-42  
spectrum, default 5-43  
Software, starting and exiting 1-3  
Sorting the peak list 3-42  
SPC file  
converting from profile to  
converting to DAT 1-30  
results, saving, opening, and  
deleting 2-40  
Settings  
see also Graphic settings  
see also Processing settings  
see also SET files  
automatically saved when data file  
closed 1-18  
Voyager, not supported 1-6  
Spec peak list, Output window 1-15  
SPEC window, see Spectrum window  
Spectra  
restoring and saving 1-20  
accumulating 4-22  
data files 5-64  
Signal intensity, summing for one  
mass 5-2  
adding/subtracting within a data  
file 4-20  
averaging 4-22  
combining (Mariner data only) 5-2,  
5-4, 7-4  
Signal-to-noise ratio  
algorithm 6-23  
calculating 6-23  
improving 5-2, 7-2, 7-3  
saturated, correcting 5-62  
single-charge conversion 5-59  
subtracting 4-20  
Index-24 Applied Biosystems  
Spectra (continued)  
summing non-contiguous 4-21  
Spectrum window (continued)  
organizing 2-13  
peak list 3-38  
E
peak, centering 2-15  
resolution, mass, calculating 6-20  
results, opening 2-40  
signal-to-noise ratio,  
calculating 6-23  
smoothing 5-42  
spectrum numbers on trace do not  
match axis 9-5  
types of 5-2  
Spectrum noise threshold, setting  
locally 3-30  
Spectrum window  
see also Peak labels  
see also Spectrum window, traces  
see also Traces  
adding spectra from different data  
files 5-64  
adding spectra from same data  
file 5-4  
baseline correction, advanced 5-48  
baseline offset 5-45  
centroid, creating 5-36  
centroiding 3-69, 5-36  
charge state labels incorrect 3-60,  
9-20  
charge state labels not  
displayed 9-19  
with 3-53  
subtracting spectra from different  
data files 5-64  
types of data 5-2  
Spectrum window, traces  
adding 2-18  
displaying as vertical bars 5-36  
do not print 2-34  
overlaying 2-25  
overlaying from different data  
files 2-24  
previewing and printing 2-33  
Spray Tip Potential, displaying  
Spreadsheet, saving peak list in 3-41  
Starting Data Explorer software 1-3  
combined spectrum, creating 5-4  
context-sensitive menus 1-14  
DAD (diode array detector),  
displaying Mariner data 4-2  
description 1-13  
display range, adjusting 2-11  
extracted absorbance  
chromatogram (XAC),  
creating from 4-13  
Stitched PSD in spectrum header 2-32,  
8-2  
Stitched spectrum, see Composite  
spectrum  
Stopping Data Explorer software 1-3  
Subtracting  
background 4-20  
raw spectra within a data file 4-20  
extracted ion chromatogram (XAC),  
creating from 4-15  
spectra, example 7-8  
extracted ion chromatogram (XIC),  
creating from 4-5, 4-8  
linking in different data files 2-13  
Mariner DAD data, displaying 1-13  
mass deconvolution 5-37  
noise filtering 5-42  
Summing data, graphic  
compression 1-28  
Summing spectra  
see Adding spectra  
see Combined spectrum  
Data Explorer Software Users Guide Index-25  
Trace labels  
T
I
spectrum 2-31  
N
D
E
X
Tabs  
for open files 2-8  
in Data Explorer window 2-8  
TAC  
Trace windows, maximizing 2-9  
Traces  
see also Trace labels  
see also Traces, copying  
see also Traces, removing  
adding 2-16, 2-18  
annotating 2-28  
axes, setting 2-11  
in chromatogram header 2-30  
Mariner data, optional 1-12  
Tag, see Event tag  
Target compounds  
determining if present in  
spectrum 6-31  
bar mode 1-28  
Technical support  
contacting 9-2  
centering a peak 2-15  
centroid, creating 5-36  
changing colors to black before  
printing 2-33  
colors, setting 1-25  
comparing 2-38  
display range, setting 2-11  
displaying as vertical bars 5-36  
distorted when you copy to another  
application 1-39  
for computers with altered  
configuration A-1  
Temperature trace, displaying 4-2  
Text files  
saving peak list in PKT 3-40  
Threshold, peak detection  
chromatogram 3-20  
global, spectrum 3-22  
local, spectrum 3-30  
dividing 2-15  
do not print 2-34  
duplicating 2-15  
expanding 2-21  
TIC  
see also Chromatogram window  
description 4-2, 4-3  
exporting to ASCII format 1-34  
filtering, see Event tag filtering  
(Mariner data only)  
graphic compression 1-28  
importing from ASCII format 1-35  
line mode 1-28  
line width 1-26  
linking 2-21  
overlaid, troubleshooting 9-6  
overlaying 2-25  
Title when saving results 2-38  
Toolbars  
buttons, adding and removing 1-21  
customizing 1-21, 1-22  
description 1-12  
Macro, displaying 6-35  
moving 1-22  
performed 5-42  
Tools commands not displayed on  
menu 9-6  
previewing 2-33  
printing 2-33  
ToolTips 1-12  
rearranging order 2-22  
recalling previously  
processed 2-22  
TR in spectrum header 2-32, 5-57  
Trace browser 2-9  
Index-26 Applied Biosystems  
Traces (continued)  
Troubleshooting (continued)  
Link View does not work 9-7  
Multiple Charge commands  
dimmed 5-37  
Replace mode, setting 2-17  
scaling mode, setting 2-12  
splitting 2-15  
switching between 2-8  
text, customizing 1-25  
traces do not print 1-26, 2-33  
type, selecting in Chromatogram  
window 4-2  
D
overlaid traces 9-6  
overview 9-2  
peak detection 9-14  
peak detection, Voyager 3-9  
peak labeling 9-14  
printer does not stay set to  
landscape mode 9-13  
printing 9-13  
use same graphic options settings  
for all 1-24  
UV, offset 4-30  
vertical bar mode 1-28  
white, does not print 1-26, 2-33  
processing 9-6  
single-charge conversion 5-62  
spectrum 9-17  
Traces, copying  
tools 9-6  
from different data file 2-37  
to Windows clipboard 1-38  
to WMF 1-39  
Truncate spectrum 5-56  
TXT file 1-34  
Traces, removing  
active 2-21  
from history list 2-22  
inactive 2-21  
Unresolved peaks, deconvoluting and  
evaluating 7-4  
Troubleshooting  
calibration 9-9  
calibration reference file (REF) not  
listed 9-10  
charge state and isotope 9-17  
charge state incorrectly  
labeled 9-17  
conversion 9-5  
Deconvolution commands  
dimmed 5-37  
elemental composition limits  
ignored 6-8, 6-11  
Unzooming 2-14  
User labels  
changing values 3-64  
creating 3-61  
customizing colors, font, size 3-61  
do not display 3-59, 3-66  
importing from other files 3-65  
saving settings 3-64  
setting values 3-64  
UV (Mariner data only)  
see also DAD data (diode array  
detector)  
error messages when you open a  
file 9-4, 9-5  
extracted ion chromatogram  
created when you  
right-click-drag to apply  
custom label 9-15  
general 9-3  
DAD data (diode array detector),  
displaying 5-2  
displaying trace 4-2  
trace offset 4-30  
Isotope calculator 6-19  
Data Explorer Software Users Guide Index-27  
Voyager Sequence Control Panel,  
automatic calibration settings  
(reference masses) for 5-28  
V
I
N
D
E
X
Valley-to-Baseline integration  
chromatogram 3-21, 3-70  
spectrum 3-26, 3-70  
VOYAGER.REF  
provided 5-18  
viewing masses in 5-10  
Valley-to-Valley integration  
chromatogram 3-21, 3-70  
spectrum 3-26, 3-70  
VOYAGERPSD.SET 8-11  
Version of software used to acquire  
data 1-15  
Vertical bars  
Warranty  
displaying centroid traces 5-36  
traces do not print 2-34  
exceptions A-3  
for computers with altered  
configuration A-1  
Vertical cursor 1-27  
Weight average molecular weight,  
determining C-15  
Vertical peak labels 3-55, 3-58  
Vial number, displaying in  
changing colors 1-20  
default settings 1-23  
SET file 1-19  
Viewing read-only files 2-7  
Views, linking 2-13  
Visual Basic Editor 6-42  
Visual Basic macros provided  
see also Data Explorer Toolbox  
accessing C-4  
activate 2-8Applied  
adding traces from same data  
file 2-16  
copying traces from different data  
files 2-37  
customizing 1-17  
displayed 1-12  
organizing 2-13, 2-36  
Ladder sequencing macro C-2  
modifying C-2  
MS Fit/MS Tag macro C-2  
overview C-2  
Peptide fragmentation macro C-2,  
C-9  
Polymer analysis macro C-2, C-15  
preparing data for C-3  
References required C-3  
Windows Metafile format, copying  
to 1-39  
Windows NT  
keywords, entering 1-31  
keywords, searching 1-32  
keywords, viewing 1-32  
Print setup 2-36  
Voyager .MS files, converting to .DAT  
not supported 1-30  
Voyager data  
WMF, copying to 1-39  
Chromatogram window,  
displaying 1-13, 2-7  
examples 7-11  
World wide web address, Applied  
Biosystems xiv  
isotope resolution limits 3-53  
peak detection strategy 3-8  
Index-28 Applied Biosystems  
X
I
X cursors, setting 1-27  
N
D
E
X
x,y data pairs, copying 1-39  
XAC  
in chromatogram header 2-30  
see also Extracted absorbance  
chromatogram (XAC)  
X-axis  
chromatogram 2-11  
setting range 2-11  
spectrum 2-11  
XIC  
see also Extracted ion  
chromatogram (XIC)  
in chromatogram header 2-31  
Y
y and b ion pairs, labeling C-11  
Y cursor, setting 1-27  
Y-axis  
offsetting chromatogram 4-27  
offsetting spectrum 5-45  
scaling 2-12  
spectrum 2-11  
Z
z labels 3-58  
z-average molecular weight,  
determining C-15  
Zero charge state displayed in peak  
list 3-40, 3-43  
Zero-charge spectrum  
for overlapping peaks 5-40  
for resolved peaks 5-37  
Zooming, multiple data files 2-13, 2-36  
Data Explorer Software Users Guide Index-29  
I
N
D
E
X
Index-30 Applied Biosystems  
Headquarters  
850 Lincoln Centre Drive  
Foster City, CA 94404 USA  
Phone: +1 650.638.5800  
Toll Free (In North America): +1 800.345.5224  
Fax: +1 650.638.5884  
Worldwide Sales and Support  
Applied Biosystems vast distribution and  
service network, composed of highly trained  
support and applications personnel, reaches  
into 150 countries on six continents. For sales  
office locations and technical support, please  
call our local office or refer to our web site at  
www.appliedbiosystems.com.  
www.appliedbiosystems.com  
Applera Corporation is committed to providing the  
world’s leading technology and information for life  
scientists. Applera Corporation consists of the  
Applied Biosystems and Celera Genomics  
businesses.  
Printed in the USA, 07/2001  
Part Number 4317717 Rev. C  

Alphasonik ALPHA 800 PSW815E User Manual
Blaupunkt BRONX MP75 User Manual
Bushnell Binoculars 111026 User Manual
Canon VIXIA HFM40 User Manual
Cardo Scala Rider MultiSet Q2 Q2 User Manual
Citizen Systems Calculator SRP 265N User Manual
Hifionics HS800 4 User Manual
Hypertec N9628PHY User Manual
Jensen Phase Linear CD1211 User Manual
JVC GR DVL308 User Manual