INSTRUCTION MANUAL
Automatic
Blood Pressure Monitor
Model BP710
Download from Www.Somanuals.com. All Manuals Search And Download.
INTRODUCTION
Thank you for purchasing the OMRON® BP710 IntelliSense® Automatic Blood
Pressure Monitor.
Fill in for future reference.
DATE PURCHASED:__________________
SERIAL NUMBER:____________________
• Staple your purchase receipt here
• Register your product on-line at
Your new digital blood pressure monitor uses the oscillometric method of blood
pressure measurement. This means the monitor detects your blood’s movement
through your brachial artery and converts the movements into a digital reading. An
oscillometric monitor does not need a stethoscope so the monitor is simple to use.
The BP710 comes with the following components:
• Monitor • Standard Adult Arm Cuff
• Instruction Manual printed in English and Spanish • Quick Start Guide
The BP710 IntelliSense® Automatic Inflation Blood Pressure Monitor is intended for
home use.
SAVE THESE INSTRUCTIONS
Download from Www.Somanuals.com3. All Manuals Search And Download.
SAfETY INfORMATION
To assure the correct use of the product, basic safety measures should always be
followed including the warnings and cautions listed in this instruction manual.
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
WARNING
Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
CAUTION
OPERATING THE DEVICE
Contact your physician for specific information about your blood pressure.
Self-diagnosis and treatment using measured results may be dangerous. Follow the
instructions of your physician or licensed healthcare provider.
The monitor is not intended to be a diagnostic device.
If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Contact a physician immediately.
Read all of the information in the instruction manual and any other literature in the
box before operating the unit.
This device is intended for use in measuring blood pressure and pulse rate in the
adult population. Do not use this device on infants or persons who cannot express
their intentions.
Read the Special Conditions section of the instruction manual if your systolic
pressure is known to be more than 220 mmHg. Inflating to a higher pressure than
necessary may result in bruising where the cuff is applied.
Operate the device only as intended. Do not use the device for any other purpose.
Dispose of the device, components and optional accessories according to applicable
local regulations. Unlawful disposal may cause environmental pollution.
Do not use a cellular phone near the device. It may result in an operational failure.
Download from Www.Somanuals.com. All Manuals Search And Download.
4
SAfETY INfORMATION
Use only Omron authorized parts and accessories. Parts and accessories not
approved for use with the device may damage the unit.
Use only 1.5 V “AAA” alkaline batteries with this device. Do not use other types of
batteries. This may damage the unit.
CARE AND MAINTENANCE
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Do not submerge the device or any of the components in water.
Store the device and the components in a clean, safe location.
Changes or modification not approved by the manufacturer will void the user
warranty. Do not disassemble or attempt to repair the unit or components.
Download from Www.Somanuals.com5. All Manuals Search And Download.
BEfORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes
before taking a measurement. Rest for at least 15 minutes before taking the
measurement.
2. Stress raises blood pressure. Avoid taking measurements during stressful times.
3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your left arm.
5. Sit on a chair with your feet flat on the floor. Rest your left arm on table so that the
cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Wait 2-3 minutes between measurements. The wait time allows the arteries to return
to the condition prior to taking the blood pressure measurement. You may need to
increase the wait time depending on your individual physiological characteristics.
8. Keep a record of your blood pressure and pulse readings for your physician. A
single measurement does not provide an accurate indication of your true blood
pressure. You need to take and record several readings over a period of time. Try to
measure your blood pressure at the same time each day for consistency.
Download from Www.Somanuals.com. All Manuals Search And Download.
6
KNOW YOUR UNIT
Main Unit:
Display
Battery compartment
START/STOP button
Memory (M) button
Air jack
Components:
INSTRUCTION MANUAL
Automatic
Blood Pressure Monitor
Air tube
Air plug
Model BP710
INTRODUCTION
®
Thank you for purchasing the OMRON BP710 Automatic Blood Pressure Monitor.
Fill in for future reference.
DATE PURCHASED: __________________
SERIAL NUMBER: ____________________
Quick Start
Guide
Instruction
Manual
Standard Adult Arm Cuff
Download from Www.Somanuals.com7. All Manuals Search And Download.
UNIT DISPLAY
Systolic blood pressure
Diastolic blood pressure
Heartbeat symbol
Pulse display
Memory symbol
(Displayed when viewing
values stored in memory)
Deflation symbol
Irregular Heartbeat symbol
Battery low indicator
Download from Www.Somanuals.com. All Manuals Search And Download.
8
DISPLAY SYMBOLS
IRREGULAR HEARTBEAT SYMBOL (
)
When the monitor detects an irregular rhythm two or more times during the
measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the
measurement values.
Normal Heartbeat
An irregular heartbeat rhythm is defined as a
rhythm that is more than 25% slower or 25%
Pulse
faster from the average rhythm detected while the
monitor is measuring the systolic and diastolic
blood pressure.
Blood
pressure
Irregular Heartbeat
If the irregular heartbeat symbol (
)
Short Long
displays with your measurement results,
we recommend you consult your
physician. Follow the directions of your
doctor.
Pulse
Blood
pressure
HEARTBEAT SYMBOL/HYPERTENSION INDICATOR (
TAKING A MEASUREMENT
)
The Heartbeat Symbol ( ) flashes on the display during the measurement. The
Heartbeat Symbol ( ) flashes at every heartbeat.
When the measurement is complete, the Heartbeat Symbol ( ) flashes on the
display with your blood pressure and pulse rate if the reading is above 135 for the
Systolic Blood Pressure value and/or above 85 for the Diastolic Blood Pressure
value.
USING THE MEMORY FUNCTION
The Heartbeat Symbol ( ) flashes when the blood pressure reading stored in the
memory is above 135 for the Systolic Blood Pressure value and/or above 85 for the
Diastolic Blood Pressure value.
Download from Www.Somanuals.com9. All Manuals Search And Download.
DISPLAY SYMBOLS
The American Heart Association recommends the following guideline for upper limit
of normal for home blood pressure readings:
AHA Home Guideline for Upper Limit of Normal BP
Systolic Blood Pressure
Diastolic Blood Pressure
135 mmHg
85 mmHg
This is a general guideline because blood pressure depends on age and morbidity.
The AHA recommends a lower home blood pressure goal for certain patients,
including diabetic patients, pregnant women, and patients with renal failure1.
WARNING
The monitor is not intended to be a diagnostic device.
WARNING
Contact your physician for specific information about your blood pressure.
Self-diagnosis and treatment using measured results may be dangerous.
Follow the instructions of your physician or licensed healthcare provider.
Omron Healthcare recommends you contact your physician if the Heartbeat Symbol
(
) flashes on the display with your measurement values.
1American Heart Association Scientific Statement 2004
Download from Www.Somanuals.com. All Manuals Search And Download.
10
BATTERY INSTALLATION
CAUTION
Use only 1.5V “AAA” alkaline batteries with this device.
1. Press the ▲ indicator on the battery cover and slide
the cover off in the direction of the arrow.
2. Install 4 “AAA” size batteries so the + (positive) and -
(negative) polarities match the polarities of the battery
compartment as indicated.
NOTE: 4 “AAA” size batteries are not included in this
unit.
3. Replace the battery cover.
BATTERY REPLACEMENT
Low Battery Indicator
When the Low Battery Indicator appears on the display, turn the monitor off and
remove all the batteries. Replace with four new batteries at the same time. Long-life
alkaline batteries are recommended.
WARNING
If battery fluid should get in your eyes, immediately rinse with plenty of
clean water. Contact a physician immediately.
CAUTION
Dispose of the device, components and optional accessories according to applicable
local regulations. Unlawful disposal may cause environmental pollution.
Download from Www.Somanuals.co1m1. All Manuals Search And Download.
APPLYING THE ARM CUff
CAUTION
Read all of the information in the instruction manual and any other literature in the
box before operating the unit.
CAUTION
This device is intended for use in measuring blood pressure and pulse rate in the
adult population. Do not use this device on infants or persons who cannot express
their intentions.
APPLYING THE CUFF ON THE LEFT ARM
1. Make sure the air plug is securely inserted in the main unit.
2. Remove tight-fitting clothing from your upper arm.
3. Apply the cuff to your left upper arm so
the arrow is centered on the inside of your
arm and aligned with your middle finger.
The air tube runs down the inside of your
arm. The bottom of the cuff should be
approximately 1/2" above your elbow.
Left Arm
1/2 inch (1 - 2 cm)
Arrow
Air Tube
Download from Www.Somanuals.com. All Manuals Search And Download.
12
APPLYING THE ARM CUff
4. Wrap the cuff firmly in place around your
arm using the cloth fastening.
APPLYING THE CUFF ON THE RIGHT ARM
When taking a measurement using the right arm, use this instruction for Step 3.
Right Arm
Apply the cuff to your right upper arm so the
arrow and the air tube runs along the inside
of your arm. The bottom of the cuff should be
approximately 1/2" above your elbow.
Arrow
NOTE: Be careful not to rest your arm on the air tube. This will restrict the flow of air
to the cuff.
Download from Www.Somanuals.co1m3. All Manuals Search And Download.
TAKING A MEASUREMENT
1. Sit on a chair with your feet flat on the floor.
Place your arm on a table so the cuff is level
with your heart.
Keep still and do not talk during measurement.
2. Press the START/STOP button.
All symbols appear on the display.
The cuff starts to inflate automatically. As the cuff inflates, the monitor
automatically determines your ideal inflation level.
Remain still and do not move your arm until the entire measurement process is completed.
NOTE: To stop the inflation or measurement, push the START/STOP button. The
monitor will stop inflating, start deflating, and will turn off.
3. Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing numbers appear on the display and the Heartbeat
Symbol ( ) flashes.
4. When the measurement is complete, the arm cuff completely
deflates. Your blood pressure and pulse rate are displayed.
Download from Www.Somanuals.com. All Manuals Search And Download.
14
TAKING A MEASUREMENT
5. Press the START/STOP button to turn the monitor off.
NOTE: The monitor will automatically turn off after five minutes.
SPECIAL CONDITIONS
CAUTION
Inflating to a higher pressure than necessary may result in bruising where the cuff is
applied.
If your systolic pressure is known to be more than 220 mmHg, press and hold the
START/STOP button until the monitor inflates 30 to 40 mmHg higher than your
expected systolic pressure.
NOTES: • Do not apply more pressure than necessary.
• The monitor will not inflate above 299 mmHg.
When the measurement is complete, the arm cuff completely deflates.
Your blood pressure and pulse rate are displayed.
Download from Www.Somanuals.co1m5. All Manuals Search And Download.
USING THE MEMORY fUNCTION
Every time you complete a measurement the monitor stores the blood pressure and pulse
rate in the memory. The monitor automatically stores up to 14 sets of measurement values
(systolic pressure, diastolic pressure, and pulse rate). When 14 sets of measurement
values are stored, the oldest record is deleted to save the most recent values.
1. Press the Memory (M) button.
The most recent set of measurement values appears on the display.
The values are displayed from the most recent to the oldest.
NOTE: The number of the reading stored in the memory appears for a second
before the pulse rate is displayed. The newest set is numbered “1”.
2. Press the START/STOP button to turn the unit off.
TO DELETE ALL THE VALUES STORED IN MEMORY
You cannot partially delete values stored in the memory. All values will be deleted.
1. Press the Memory (M) button.
Then while holding it down, press the START/STOP button down for more than 2
seconds until the display appears as illustrated below.
2. Press the START/STOP button to turn the unit off.
NOTE: The monitor will automatically turn off after five minutes.
Download from Www.Somanuals.com. All Manuals Search And Download.
16
CARE AND MAINTENANCE
To keep your digital blood pressure monitor in the best condition and protect the unit
from damage, follow the directions listed below:
Do not forcefully bend the arm cuff or air tube. Do not fold tightly.
Clean the monitor with a soft dry cloth.
Do not use any abrasive or volatile cleaners.
Do not attempt to clean the cuff.
CAUTION
Do not submerge the device or any of the components in water. Do
not subject the monitor to extreme hot or cold temperatures,
humidity or direct sunlight.
CAUTION
Store the device and the components in a clean, safe location.
CAUTION
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Remove the batteries if the unit will not be used for three months or longer. Always
replace all the batteries with new ones at the same time.
Use the unit consistent with the instruction provided in this manual.
Use only authorized parts and accessories.
Use the unit in a manner consistent with the instructions provided in this manual.
CAUTION
Use only Omron authorized parts and accessories. Parts and accessories not
approved for use with the device may damage the unit.
CAUTION
Changes or modification not approved by the manufacturer will void the user
warranty. Do not disassemble or attempt to repair the unit or components.
Download from Www.Somanuals.co1m7. All Manuals Search And Download.
ERROR INDICATORS
SYMBOL
CAUSE
CORRECTION
Take another measurement and
remain still until the measurement
is complete.
Monitor could not
detect pulse wave.
Cuff under-inflated.
Cuff not applied
correctly.
Remove the arm cuff.
Read “Taking a Measurement”.
Wait 2-3 minutes.
Take another measurement.
Cuff over-inflated.
Batteries are worn.
Device error.
Replace the four batteries. Refer to
“Battery Installation”.
Contact Customer Service.
Download from Www.Somanuals.com. All Manuals Search And Download.
18
TROUBLESHOOTING TIPS
PROBLEM
CAUSES AND SOLUTIONS
No power.
Replace all four batteries with new ones.
No display appears on the unit. Check the battery installation for proper placement
of the battery polarities.
Blood pressure varies constantly. Many factors
including stress, time of day, how you wrap the
cuff, may affect your blood pressure.
Review the sections “Before Taking a Measure-
Measurement values appear too
high or too low.
ment” and “Taking a Measurement”.
Download from Www.Somanuals.co1m9. All Manuals Search And Download.
fCC STATEMENT
Note:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference
in a residential installation. The product generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If the product does
cause harmful interference to radio or television reception, which can be determined by
turning the product on and off, the user is encouraged to try to correct the interference
by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and the receiver.
• Connect the product into an outlet on a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise emissions
from digital apparatus as set out in the interference-causing equipment standard entitled
“Digital Apparatus”, ICES-003 of the Canadian Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux
appareils numériques de Clase B prescrites dans la norme sur le materiel brouilleur:
“Appareils Numériques”, ICES-003 édictée par le minister des communications.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
Download from Www.Somanuals.com. All Manuals Search And Download.
20
LIMITED WARRANTIES
Your BP710 IntelliSense® Automatic Blood Pressure Monitor, excluding the arm cuff,
is warranted to be free from defects in materials and workmanship appearing within
2 years from the date of purchase, when used in accordance with the instructions
provided with the monitor. The arm cuff is warranted to be free from defects in
materials and workmanship appearing within one year from the date of purchase when
the monitor is used in accordance with the instructions provided with the monitor. The
above warranties extend only to the original retail purchaser.
We will, at our option, repair or replace without charge any monitor or arm cuff
covered by the above warranties. Repair or replacement is our only responsibility and
your only remedy under the above warranties.
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the
address of the repair location and the return shipping and handling fee.
Enclose the Proof of Purchase. Include a letter, with your name, address, phone
number, and description of the specific problem. Pack the product carefully to prevent
damage in transit. Because of possible loss in transit, we recommend insuring the
product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS
ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS
THAT MAY BE IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION
TO THE PERIOD OF THE ABOVE EXPRESS WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTAIL OR INDIRECT COSTS,
EXPENSES OR DAMAGES.
This warranty provides you with specific legal rights, and you may have other rights
that vary by jurisdiction. Because of special local requirements, some of the above
limitations and exclusions may not apply to you.
FOR CUSTOMER SERVICE
Visit our web site at:
Call toll free:
1-800-634-4350
Download from Www.Somanuals.co2m1. All Manuals Search And Download.
SPECIfICATIONS
BP710 (HEM-7113-Z)
LCD Digital Display
Model:
Display:
Pressure: 0 to 299 mmHg,
Pulse: 40 to 180/min.
Measurement Range:
Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading
Accuracy/Calibration:
Automatic by electric pump
Inflation:
Automatic pressure release valve
Automatic exhaust valve
Deflation:
Rapid Air Release:
Measurement Method:
Power Source:
Battery Life:
Oscillometric method
1.5V 4 “AAA” batteries
Approx. 300 uses with 4 new alkaline batteries
50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH
Operating Temperature /
Humidity:
-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
700-1060 hPa
Storage Temperature /
Humidity/Air Pressure:
Approximately 8 5/8 oz. (245g)
not including batteries
Main Unit Weight:
Main Unit Dimensions:
Cuff Size:
Approx. 5"(l) × 4 1/8"(w) × 2 1/2" (h)
(128 mm × 104 mm × 64 mm)
Approximately 5 3/4"(w) × 17 1/2"(l) (146 mm × 446 mm)
Cuff tube 23 5/8" (600 mm)
Fits arm circumferences 9" to 13"
(220 mm to 320 mm)
Cuff Circumference:
Download from Www.Somanuals.com. All Manuals Search And Download.
22
SPECIfICATIONS
Up to 14 readings
Memory:
Contents:
Main Unit, Standard Adult Arm Cuff, Instruction
Manual, and Quick Start Guide
Large Cuff 13" - 17" (320 mm - 420 mm)
Optional Accessories:
= Type B
NOTE: These specifications are subject to change without notice.