SECTION 1. INTRODUCTION  
1.1 UNPACKING  
Remove the Packing List and verify that all equipment has been received. If  
there are any questions about the shipment, use the phone numbers listed on  
the back cover to contact the Customer Service Department nearest you.  
Upon receipt of shipment, inspect the container and equipment for any signs of  
damage. Take particular note of any evidence of rough handling in transit.  
Immediately report any damage to the shipping agent.  
The carrier will not honor any claims unless all shipping material  
is saved for their examination. After examining and removing  
contents, save packing material and carton in the event  
reshipment is necessary.  
Verify that you receive the following items in the shipping box:  
QTY  
DESCRIPTION  
1
Differential THermocouple Meter with all applicable connectors  
attached.  
1
1
1
Owner's Manual  
Set Mounting brackets  
QuickStart Manual  
If you ordered any of the available options (except the "BL"  
Blank Lens option), they will be shipped in a separate container  
to avoid any damage to your indicator/controller.  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.2 SAFETY CONSIDERATIONS  
This device is marked with the international caution symbol. It is important  
to read this manual before installing or commissioning this device as it contains  
important information relating to Safety and EMC (Electromagnetic  
Compatibility).  
This instrument is a panel mount device protected in accordance with EN  
61010-1:2001, electrical safety requirements for electrical equipment for  
measurement, control and laboratory. Installation of this instrument should be  
done by qualified personnel. In order to ensure safe operation, the following  
instructions should be followed.  
This instrument has no power-on switch. An external switch or circuit-  
breaker shall be included in the building installation as a disconnecting device.  
It shall be marked to indicate this function, and it shall be in close proximity to  
the equipment within easy reach of the operator. The switch or circuit-breaker  
shall not interrupt the Protective Conductor (Earth wire), and it shall meet the  
relevant requirements of IEC 947–1 and IEC 947-3 (International  
Electrotechnical Commission). The switch shall not be incorporated in the main  
supply cord.  
Furthermore, to provide protection against excessive energy being drawn  
from the main supply in case of a fault in the equipment, an overcurrent  
protection device shall be installed.  
Do not exceed voltage rating on the label located on the top of the  
instrument housing.  
Always disconnect power before changing signal and power connections.  
Do not use this instrument on a work bench without its case for safety  
reasons.  
Do not operate this instrument in flammable or explosive atmospheres.  
Do not expose this instrument to rain or moisture.  
Unit mounting should allow for adequate ventilation to ensure instrument  
does not exceed operating temperature rating.  
Use electrical wires with adequate size to handle mechanical strain and  
power requirements. Install without exposing bare wire outside the  
connector to minimize electrical shock hazards.  
EMC Considerations  
Whenever EMC is an issue, always use shielded cables.  
Never run signal and power wires in the same conduit.  
Use signal wire connections with twisted-pair cables.  
Install Ferrite Bead(s) on signal wires close to the instrument if EMC  
problems persist.  
Failure to follow all instructions and warnings may result in injury!  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 2. ABOUT THE METER  
2.1 DESCRIPTION  
The Differential Temperature meter with thermocouple is a value packed  
indicator/ controller. Four full digits allow for an accurate display of your  
temperature. Select from J, K, T or DIN J thermocouple types. Your  
meter may be a basic indicator or it may include analog output or dual  
relay output. Analog or dual relay output must be ordered at time of  
purchase. Analog output is fully scalable and may be configured as a  
proportional controller, or to follow your display. Dual 5 amp, form C  
relays control critical processes. Front panel peak detection and memory  
is also standard.  
2.2 STANDARD FEATURES  
The following is a list of features:  
*
*
*
*
*
*
*
*
4-digit, red, 14 segment LED Display  
NEMA 4/Type 4 Front Bezel  
0.5 °C accuracy  
Cross channel isolation up to 5V  
Match Capability  
Automatic Decimal Adjustment  
Non-volatile memory-no battery backup  
115 or 230 Vac 50/60 Hz power supply or  
10-32 Vdc or 26-56 Vdc  
*
T1, T2, T1-T2 display  
OPTIONAL FEATURES  
*
*
*
*
*
Dual 5 amp, form C relay outputs  
Scalable analog output  
Analog out proportional control  
Front-panel deviation correction  
Easy setup for proportional control  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.5 BACK OF THE METER  
Figure 2-2 shows the connector label mounted at the top of the meter  
housing. Table 2-2 gives a brief description of each connector at the back  
of the meter.  
Figure 2-2 Connector Label (ac power with dc detail)  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.3 PANEL MOUNTING  
PANEL  
CUT-OUT  
"NEW" STYLE  
MOUNTING  
BRACKET  
CONNECTOR  
LABEL  
PRODUCT  
LABEL  
CASE  
"OLDER" STYLE  
MOUNTING  
BRACKET 2 PCS  
GASKET  
FRONT BEZEL  
Figure 3-3 Meter - Exploded VIew  
1. Cut a hole in your  
panel, as shown in  
Figure 3-3. For  
specific  
dimensions refer to  
Figure 3-4.  
2. Insert the meter  
into the hole.  
Be sure the front  
bezel gasket is  
flush to the panel.  
3. Slide on mounting  
bracket to secure.  
4. Proceed to Section  
3.4 to connect your  
sensor input and  
main power.  
Figure 3-4 Panel Cut-Out  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 13. SPECIFICATIONS (continued)  
ALARM OUTPUTS  
(if applicable)  
2 Form "C" on/off relays. Configurable for latched and  
unlatched by software.  
Max current: 5 AMPS, resistive load  
Max voltage: 250 Vac or 30 Vdc  
ENVIRONMENT  
~
AC units  
115/230 V~(AC) + 10%, 50/60 Hz  
7 W max, power consumption (Non-Isolated Analog Out)  
8 W max, power consumption (Isolated Analog Out)  
DC units  
10-32 Vdc or 26-56 Vdc,  
6 W max, power consumption (Non-Isolated Analog Out)  
7 W max, power consumption (Isolated Analog Out)  
External Fuse Required:  
IEC 127-2/III  
Power  
115 V  
230 V  
Fuse  
125 mA @ 250 (T)  
63 mA @ 250 (T)  
UL 248-14 (Listed Fuse)  
Power  
115 V  
230 V  
Fuse  
175 mA @ 250 V Slow-Blow  
80 mA @ 250 V Slow-Blow  
ENVIRONMENT  
Operating temperature:  
0° to 50°C (32° to 122°F)  
(115/230 V rms + 10%)  
Storage temperature:  
Relative humidity:  
-40° to 85°C (-40° to 185° F)  
90% at 40°C (non-condensing)  
MECHANICAL  
Panel cutout:  
Weight:  
1/8 DIN 3.62 x 1.78" (45 x 92mm)  
1.27 lb (575 g)  
Case material:  
Protection:  
Polycarbonate, 94 V-O UL rated  
NEMA-4/Type 4 Front BezeL  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 13. SPECIFICATIONS (continued)  
48,0 (1.89)  
96,0 (3.78)  
FRONT BEZEL  
20,3 (.80)  
RETAINER  
CASE  
REAR COVER  
TOP VIEW  
SIDE VIEW  
Figure 13-1 Meter Dimensions  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 14. FACTORY PRESET VALUES  
Table 14-1. Factory Preset Values  
MENU ITEM  
FACTORY PRESET VALUES  
INPT  
Input Type: K.TC (Type K T/C)  
Decimal Point Position: FFF.F  
Reading Configuration:  
DEC.P  
RD.CF  
R.1=F (Fahrenheit)  
R.2=D (Shows deviation in “RUN” mode)  
Setpoint 1 Configuration:  
S1.CF  
S2.CF  
S.1=A (Setpoint is active above)  
S.2=U (Setpoint is unlatched)  
S.3=0 (Setpoint 1 assigned to T1 - T2)  
Setpoint 2 Configuration:  
S.1=A (Setpoint is active above)  
S.2=U (Setpoint is unlatched)  
S.3=0 (Setpoint 2 assigned to T1 - T2)  
Setpoint 1 Deadband: 3.0  
S1.DB  
S2.DB  
OT.CF  
Setpoint 2 Deadband: 3.0  
Output Configuration:  
O.1=E (Analog output is enabled)  
O.2=C (Analog output is current)  
O.3=A (Analog output follows the display value)  
Analog Output Routing: T1 - T2  
Output Scale and Offset:  
OUT  
OT.S.O  
0-1000 = 4-20 mA dc  
C.OF.1  
Cold Junction Offset for T1: 000.0  
Cold Junction Offset for T2: 000.0  
Setpoint 1 Value: 000.0  
SP1  
SP2  
Setpoint 2 Value: 000.0  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CE APPROVALS INFORMATION  
This product conforms to the EMC directive 89/336/EEC amended by  
93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.  
Electrical Safety EN61010-1:2001  
Safety requirements for electrical equipment for measurement, control and  
laboratory.  
Double Insulation  
Pollution Degree 2  
Dielectric withstand Test per 1 min  
Power to Input/Output:  
Power to Input/Output:  
(Low Voltage dc Power Option*)  
2300 Vac (3250 Vdc)  
500 Vac (720 Vdc)  
Power to Relays Output:  
Relay 1 to Relay 2:  
Isolated Analog to Inputs:  
Analog to Inputs:  
2300 Vac (3250 Vdc)  
2300 Vac (3250 Vdc)  
1000 Vac (1420 Vdc)  
No Isolation  
Measurement Category I  
Category I are measurements performed on circuits not directly connected to the  
Mains Supply (power). Maximum Line-to-Neutral working voltage is 50 Vac/dc.  
This unit should not be used in Measurement Categories II, III, IV.  
Transients Overvoltage Surge (1.2 / 50uS pulse)  
Input Power:  
Input Power:  
(Low Voltage dc Power Option*)  
2500 V  
500 V  
Isolated Analog:  
Input/Output Signals:  
500 V  
500 V  
Note: *Units configured for external low power dc voltage, 10-32 Vdc (Basic Insulation)  
EMC EN61326:1997 + and A1:1998 + A2:2001  
Immunity and Emissions requirements for electrical equipment for measurement,  
control and laboratory.  
EMC Emissions Table 4, Class B of EN61326  
EMC Immunity** Table 1 of EN61326  
Note: **I/O signal and control lines require shielded cables and these cables  
must be located on conductive cable trays or in conduits. Furthermore,  
the length of these cables should not exceed 30 meters  
Refer to the EMC and Safety installation considerations (Guidelines) of this manual  
for additional information.  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARRANTY/DISCLAIMER  
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of  
one (1) year from the date of purchase. In addition to OMEGA’s standard warranty period, OMEGA Engineering will  
extend the warranty period for four (4) additional years if the warranty card enclosed with each instrument is  
returned to OMEGA.  
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department  
will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by  
OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY  
does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling,  
improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This  
WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been  
damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification;  
misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not  
warranted, including but not limited to contact points, fuses, and triacs.  
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither  
assumes responsibility for any omissions or errors nor assumes liability for any damages that result from  
the use of its products in accordance with information provided by OMEGA, either verbal or written.  
OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA  
MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR  
IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF  
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.  
LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability  
of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification,  
strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is  
based. In no event shall OMEGA be liable for consequential, incidental or special damages.  
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic  
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical  
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity,  
medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth  
in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and  
hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in  
such a manner.  
RETURN REQUESTS/INQUIRIES  
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE  
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN  
(AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID  
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return  
package and on any correspondence.  
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent  
breakage in transit.  
FOR WARRANTY RETURNS, please have the  
following information available BEFORE  
contacting OMEGA:  
FOR NON-WARRANTY REPAIRS, consult OMEGA for  
current repair charges. Have the following information  
available BEFORE contacting OMEGA:  
1. Purchase Order number under which the product  
was PURCHASED,  
1. Purchase Order number to cover the COST  
of the repair,  
2. Model and serial number of the product under  
warranty, and  
3. Repair instructions and/or specific problems  
relative to the product.  
2. Model and serial number of product, and  
3. Repair instructions and/or specific problems  
relative to the product.  
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our  
customers the latest in technology and engineering.  
© Copyright 2007 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced,  
translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of  
OMEGA ENGINEERING, INC.  
®
®
®
TRADEMARK NOTICE:  
and  
are trademarks of OMEGA Engineering, Inc.  
omega.com  
,
PATENT NOTICE: This product is covered by one or more of the following patents: U.S. Pat. No. Des. 336,895; 5,274,577; 6,243,021  
/ CANADA 2052599; 2052600 / ITALY 1249456; 1250938 / FRANCE BREVET No. 91 12756 / SPAIN 2039150; 2048066 /  
UK PATENT No. GB2 249 837; GB2 248 954 / GERMANY DE 41 34398 C2. The “Meter Case Bezel Design” is a trademark of  
NEWPORT Electronics, Inc., registered in the U.S. USED UNDER LICENSE. Other US and International Patents pending or  
applied for.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Where Do I Find Everything I Need for  
Process Measurement and Control?  
OMEGA…Of Course!  
Shop on line at omega.com  
TEMPERATURE  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies  
Wire: Thermocouple, RTD & Thermistor  
Calibrators & Ice Point References  
Recorders, Controllers & Process Monitors  
Infrared Pyrometers  
PRESSURE, STRAIN AND FORCE  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Transducers & Strain Gauges  
Load Cells & Pressure Gauges  
Displacement Transducers  
Instrumentation & Accessories  
FLOW/LEVEL  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Rotameters, Gas Mass Flowmeters & Flow Computers  
Air Velocity Indicators  
Turbine/Paddlewheel Systems  
Totalizers & Batch Controllers  
pH/CONDUCTIVITY  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
pH Electrodes, Testers & Accessories  
Benchtop/Laboratory Meters  
Controllers, Calibrators, Simulators & Pumps  
Industrial pH & Conductivity Equipment  
DATA ACQUISITION  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Data Acquisition & Engineering Software  
Communications-Based Acquisition Systems  
Plug-in Cards for Apple, IBM & Compatibles  
Datalogging Systems  
Recorders, Printers & Plotters  
HEATERS  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Heating Cable  
Cartridge & Strip Heaters  
Immersion & Band Heaters  
Flexible Heaters  
Laboratory Heaters  
ENVIRONMENTAL  
MONITORING AND CONTROL  
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
ߜ
 
Metering & Control Instrumentation  
Refractometers  
Pumps & Tubing  
Air, Soil & Water Monitors  
Industrial Water & Wastewater Treatment  
pH, Conductivity & Dissolved Oxygen Instruments  
11797ML-97 Rev C  
M1677/0307  
Download from Www.Somanuals.com. All Manuals Search And Download.  

MTD Snow Blower 199 658 User Manual
Multiquip Chainsaw QP 301TH User Manual
Mustang Survival Life Jacket md2010 User Manual
Nilfisk Advance America Vacuum Cleaner BRV 900 User Manual
Nortel Networks Cordless Telephone IP Phone 1230 User Manual
Nostalgia Electrics Popcorn Poppers MHP 310 User Manual
Omron Blood Pressure Monitor BP762 User Manual
OWI Portable Speaker LGS300DVC User Manual
Panasonic All in One Printer KX FC175AL User Manual
Panasonic Car Stereo System CQ BT5107U User Manual