SECTION 1. INTRODUCTION
1.1 UNPACKING
Remove the Packing List and verify that all equipment has been received. If
there are any questions about the shipment, use the phone numbers listed on
the back cover to contact the Customer Service Department nearest you.
Upon receipt of shipment, inspect the container and equipment for any signs of
damage. Take particular note of any evidence of rough handling in transit.
Immediately report any damage to the shipping agent.
The carrier will not honor any claims unless all shipping material
is saved for their examination. After examining and removing
contents, save packing material and carton in the event
reshipment is necessary.
Verify that you receive the following items in the shipping box:
QTY
DESCRIPTION
1
Differential THermocouple Meter with all applicable connectors
attached.
1
1
1
Owner's Manual
Set Mounting brackets
QuickStart Manual
If you ordered any of the available options (except the "BL"
Blank Lens option), they will be shipped in a separate container
to avoid any damage to your indicator/controller.
1
Download from Www.Somanuals.com. All Manuals Search And Download.
1.2 SAFETY CONSIDERATIONS
This device is marked with the international caution symbol. It is important
to read this manual before installing or commissioning this device as it contains
important information relating to Safety and EMC (Electromagnetic
Compatibility).
This instrument is a panel mount device protected in accordance with EN
61010-1:2001, electrical safety requirements for electrical equipment for
measurement, control and laboratory. Installation of this instrument should be
done by qualified personnel. In order to ensure safe operation, the following
instructions should be followed.
This instrument has no power-on switch. An external switch or circuit-
breaker shall be included in the building installation as a disconnecting device.
It shall be marked to indicate this function, and it shall be in close proximity to
the equipment within easy reach of the operator. The switch or circuit-breaker
shall not interrupt the Protective Conductor (Earth wire), and it shall meet the
relevant requirements of IEC 947–1 and IEC 947-3 (International
Electrotechnical Commission). The switch shall not be incorporated in the main
supply cord.
Furthermore, to provide protection against excessive energy being drawn
from the main supply in case of a fault in the equipment, an overcurrent
protection device shall be installed.
•
Do not exceed voltage rating on the label located on the top of the
instrument housing.
•
•
Always disconnect power before changing signal and power connections.
Do not use this instrument on a work bench without its case for safety
reasons.
•
•
•
Do not operate this instrument in flammable or explosive atmospheres.
Do not expose this instrument to rain or moisture.
Unit mounting should allow for adequate ventilation to ensure instrument
does not exceed operating temperature rating.
•
Use electrical wires with adequate size to handle mechanical strain and
power requirements. Install without exposing bare wire outside the
connector to minimize electrical shock hazards.
EMC Considerations
•
•
•
•
Whenever EMC is an issue, always use shielded cables.
Never run signal and power wires in the same conduit.
Use signal wire connections with twisted-pair cables.
Install Ferrite Bead(s) on signal wires close to the instrument if EMC
problems persist.
Failure to follow all instructions and warnings may result in injury!
2
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 2. ABOUT THE METER
2.1 DESCRIPTION
The Differential Temperature meter with thermocouple is a value packed
indicator/ controller. Four full digits allow for an accurate display of your
temperature. Select from J, K, T or DIN J thermocouple types. Your
meter may be a basic indicator or it may include analog output or dual
relay output. Analog or dual relay output must be ordered at time of
purchase. Analog output is fully scalable and may be configured as a
proportional controller, or to follow your display. Dual 5 amp, form C
relays control critical processes. Front panel peak detection and memory
is also standard.
2.2 STANDARD FEATURES
The following is a list of features:
*
*
*
*
*
*
*
*
4-digit, red, 14 segment LED Display
NEMA 4/Type 4 Front Bezel
0.5 °C accuracy
Cross channel isolation up to 5V
Match Capability
Automatic Decimal Adjustment
Non-volatile memory-no battery backup
115 or 230 Vac 50/60 Hz power supply or
10-32 Vdc or 26-56 Vdc
*
T1, T2, T1-T2 display
OPTIONAL FEATURES
*
*
*
*
*
Dual 5 amp, form C relay outputs
Scalable analog output
Analog out proportional control
Front-panel deviation correction
Easy setup for proportional control
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
2.5 BACK OF THE METER
Figure 2-2 shows the connector label mounted at the top of the meter
housing. Table 2-2 gives a brief description of each connector at the back
of the meter.
Figure 2-2 Connector Label (ac power with dc detail)
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
3.3 PANEL MOUNTING
PANEL
CUT-OUT
"NEW" STYLE
MOUNTING
BRACKET
CONNECTOR
LABEL
PRODUCT
LABEL
CASE
"OLDER" STYLE
MOUNTING
BRACKET 2 PCS
GASKET
FRONT BEZEL
Figure 3-3 Meter - Exploded VIew
1. Cut a hole in your
panel, as shown in
Figure 3-3. For
specific
dimensions refer to
Figure 3-4.
2. Insert the meter
into the hole.
Be sure the front
bezel gasket is
flush to the panel.
3. Slide on mounting
bracket to secure.
4. Proceed to Section
3.4 to connect your
sensor input and
main power.
Figure 3-4 Panel Cut-Out
15
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 13. SPECIFICATIONS (continued)
ALARM OUTPUTS
(if applicable)
2 Form "C" on/off relays. Configurable for latched and
unlatched by software.
Max current: 5 AMPS, resistive load
Max voltage: 250 Vac or 30 Vdc
ENVIRONMENT
~
AC units
115/230 V~(AC) + 10%, 50/60 Hz
7 W max, power consumption (Non-Isolated Analog Out)
8 W max, power consumption (Isolated Analog Out)
DC units
10-32 Vdc or 26-56 Vdc,
6 W max, power consumption (Non-Isolated Analog Out)
7 W max, power consumption (Isolated Analog Out)
External Fuse Required:
IEC 127-2/III
Power
115 V
230 V
Fuse
125 mA @ 250 (T)
63 mA @ 250 (T)
UL 248-14 (Listed Fuse)
Power
115 V
230 V
Fuse
175 mA @ 250 V Slow-Blow
80 mA @ 250 V Slow-Blow
ENVIRONMENT
Operating temperature:
0° to 50°C (32° to 122°F)
(115/230 V rms + 10%)
Storage temperature:
Relative humidity:
-40° to 85°C (-40° to 185° F)
90% at 40°C (non-condensing)
MECHANICAL
Panel cutout:
Weight:
1/8 DIN 3.62 x 1.78" (45 x 92mm)
1.27 lb (575 g)
Case material:
Protection:
Polycarbonate, 94 V-O UL rated
NEMA-4/Type 4 Front BezeL
49
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 13. SPECIFICATIONS (continued)
48,0 (1.89)
96,0 (3.78)
FRONT BEZEL
20,3 (.80)
RETAINER
CASE
REAR COVER
TOP VIEW
SIDE VIEW
Figure 13-1 Meter Dimensions
50
Download from Www.Somanuals.com. All Manuals Search And Download.
SECTION 14. FACTORY PRESET VALUES
Table 14-1. Factory Preset Values
MENU ITEM
FACTORY PRESET VALUES
INPT
Input Type: K.TC (Type K T/C)
Decimal Point Position: FFF.F
Reading Configuration:
DEC.P
RD.CF
R.1=F (Fahrenheit)
R.2=D (Shows deviation in “RUN” mode)
Setpoint 1 Configuration:
S1.CF
S2.CF
S.1=A (Setpoint is active above)
S.2=U (Setpoint is unlatched)
S.3=0 (Setpoint 1 assigned to T1 - T2)
Setpoint 2 Configuration:
S.1=A (Setpoint is active above)
S.2=U (Setpoint is unlatched)
S.3=0 (Setpoint 2 assigned to T1 - T2)
Setpoint 1 Deadband: 3.0
S1.DB
S2.DB
OT.CF
Setpoint 2 Deadband: 3.0
Output Configuration:
O.1=E (Analog output is enabled)
O.2=C (Analog output is current)
O.3=A (Analog output follows the display value)
Analog Output Routing: T1 - T2
Output Scale and Offset:
OUT
OT.S.O
0-1000 = 4-20 mA dc
C.OF.1
Cold Junction Offset for T1: 000.0
Cold Junction Offset for T2: 000.0
Setpoint 1 Value: 000.0
SP1
SP2
Setpoint 2 Value: 000.0
51
Download from Www.Somanuals.com. All Manuals Search And Download.
CE APPROVALS INFORMATION
This product conforms to the EMC directive 89/336/EEC amended by
93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and
laboratory.
Double Insulation
Pollution Degree 2
Dielectric withstand Test per 1 min
• Power to Input/Output:
• Power to Input/Output:
(Low Voltage dc Power Option*)
2300 Vac (3250 Vdc)
500 Vac (720 Vdc)
• Power to Relays Output:
• Relay 1 to Relay 2:
• Isolated Analog to Inputs:
• Analog to Inputs:
2300 Vac (3250 Vdc)
2300 Vac (3250 Vdc)
1000 Vac (1420 Vdc)
No Isolation
Measurement Category I
Category I are measurements performed on circuits not directly connected to the
Mains Supply (power). Maximum Line-to-Neutral working voltage is 50 Vac/dc.
This unit should not be used in Measurement Categories II, III, IV.
Transients Overvoltage Surge (1.2 / 50uS pulse)
• Input Power:
• Input Power:
(Low Voltage dc Power Option*)
2500 V
500 V
• Isolated Analog:
• Input/Output Signals:
500 V
500 V
Note: *Units configured for external low power dc voltage, 10-32 Vdc (Basic Insulation)
EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement,
control and laboratory.
• EMC Emissions Table 4, Class B of EN61326
• EMC Immunity** Table 1 of EN61326
Note: **I/O signal and control lines require shielded cables and these cables
must be located on conductive cable trays or in conduits. Furthermore,
the length of these cables should not exceed 30 meters
Refer to the EMC and Safety installation considerations (Guidelines) of this manual
for additional information.
52
Download from Www.Somanuals.com. All Manuals Search And Download.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of
one (1) year from the date of purchase. In addition to OMEGA’s standard warranty period, OMEGA Engineering will
extend the warranty period for four (4) additional years if the warranty card enclosed with each instrument is
returned to OMEGA.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department
will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by
OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY
does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling,
improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This
WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been
damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification;
misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not
warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither
assumes responsibility for any omissions or errors nor assumes liability for any damages that result from
the use of its products in accordance with information provided by OMEGA, either verbal or written.
OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA
MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR
IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability
of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification,
strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is
based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity,
medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth
in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and
hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in
such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN
(AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return
package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR WARRANTY RETURNS, please have the
following information available BEFORE
contacting OMEGA:
FOR NON-WARRANTY REPAIRS, consult OMEGA for
current repair charges. Have the following information
available BEFORE contacting OMEGA:
1. Purchase Order number under which the product
was PURCHASED,
1. Purchase Order number to cover the COST
of the repair,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
2. Model and serial number of product, and
3. Repair instructions and/or specific problems
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our
customers the latest in technology and engineering.
© Copyright 2007 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced,
translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of
OMEGA ENGINEERING, INC.
®
®
®
TRADEMARK NOTICE:
and
are trademarks of OMEGA Engineering, Inc.
omega.com
,
PATENT NOTICE: This product is covered by one or more of the following patents: U.S. Pat. No. Des. 336,895; 5,274,577; 6,243,021
/ CANADA 2052599; 2052600 / ITALY 1249456; 1250938 / FRANCE BREVET No. 91 12756 / SPAIN 2039150; 2048066 /
UK PATENT No. GB2 249 837; GB2 248 954 / GERMANY DE 41 34398 C2. The “Meter Case Bezel Design” is a trademark of
NEWPORT Electronics, Inc., registered in the U.S. USED UNDER LICENSE. Other US and International Patents pending or
applied for.
Download from Www.Somanuals.com. All Manuals Search And Download.
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop on line at omega.com
TEMPERATURE
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
Wire: Thermocouple, RTD & Thermistor
Calibrators & Ice Point References
Recorders, Controllers & Process Monitors
Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Transducers & Strain Gauges
Load Cells & Pressure Gauges
Displacement Transducers
Instrumentation & Accessories
FLOW/LEVEL
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Rotameters, Gas Mass Flowmeters & Flow Computers
Air Velocity Indicators
Turbine/Paddlewheel Systems
Totalizers & Batch Controllers
pH/CONDUCTIVITY
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ pH Electrodes, Testers & Accessories
Benchtop/Laboratory Meters
Controllers, Calibrators, Simulators & Pumps
Industrial pH & Conductivity Equipment
DATA ACQUISITION
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Data Acquisition & Engineering Software
Communications-Based Acquisition Systems
Plug-in Cards for Apple, IBM & Compatibles
Datalogging Systems
Recorders, Printers & Plotters
HEATERS
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Heating Cable
Cartridge & Strip Heaters
Immersion & Band Heaters
Flexible Heaters
Laboratory Heaters
ENVIRONMENTAL
MONITORING AND CONTROL
Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Ⅺ
ߜ Metering & Control Instrumentation
Refractometers
Pumps & Tubing
Air, Soil & Water Monitors
Industrial Water & Wastewater Treatment
pH, Conductivity & Dissolved Oxygen Instruments
11797ML-97 Rev C
M1677/0307
Download from Www.Somanuals.com. All Manuals Search And Download.
|