1 About this operating manual  
The operating manual is aimed at specialists and semi-skilled personnel.  
Before each step, read through the relevant advice carefully and keep to the specified order.  
Thoroughly read and understand the information in the section (Safety Instructions "3").  
If you have any problems or questions, please contact:  
One Omega Dr. Box 4047  
Stamford, CT 06907-0047  
Tel: (203) 359-1660  
Hazard signs and other symbols used:  
WARNING! / CAUTION! Risk of injury!  
This sign indicates dangers that cause personal injuries that can lead to health defects or  
cause considerable damage to property.  
CAUTION! Electric current!  
This sign indicates dangers which could arise from handling of electric current.  
CAUTION! Material damage!  
This sign indicates actions which could lead to possible damage to material or environmental  
damage.  
NOTICE!  
ADHERE TO OPERATING MANUAL!  
This symbol indicates important notices,  
tips or information.  
NO DOMESTIC WASTE!  
The device must not be disposed of  
together with domestic waste.  
Check the specified points or notices.  
Pay attention to and comply with  
information that is marked with this symbol.  
J Reference to another section, document or  
source.  
ª Follow the specified instructions and steps.  
Adhere to the given order.  
Item.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
2 Device description  
The FMG90 series from Omega, is a non-contact flow sensor. The measurement is  
performed using magnetic induction and works without any moving parts.  
The FMG90 is used for measuring or metering water and aqueous solutions. The compact  
design and independence from the intake and discharge sections allows the FMG90 to be  
used under a variety of conditions.  
Components:  
c Sensor housing:  
The sensor housing consists of plastic and has the IP65 degree of protection.  
d Electrical connection:  
The electrical connection is made via 4-pin plug M12x1.  
e Operation / flow indicator LED.  
f Type plate with flow direction (marking)  
g Process connection:  
The process connections are available in different sizes.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
2.1 Intended use  
The magnetic inductive flow sensor FMG90 must only be used for measuring and metering  
liquids with a minimum conductivity of 20 μS/cm.  
WARNING! No safety component!  
The magnetic inductive flow sensor of the series FMG90 are not safety components in  
accordance with Directive 2006-42-EC (Machine Directive).  
ª Never use the FMG90 as a safety component.  
The operational safety of the device supplied is only guaranteed within its intended use. The  
specified limits (see Technical Data "10") may under no circumstances be exceeded.  
Before installing the device, check that the wetted materials of the device are compatible with  
the media being used (see Materials Table "10.1").  
Measuring tube empty (or partially filled). / Conductivity too low.  
The green LED may blink irregularly if the measuring tube of the FMG90 is empty or  
partially filled or if the conductivity of the fluid being used is too low. Random pulses will  
be present at the output, but they do not represent an actual flow.  
ª Ensure that the measuring tube of the FMG90 is always completely filled (see  
Installation Instructions "5.1").  
ª Ensure that the conductivity of the fluid is at least 20 μS/cm.  
2.1 Exclusion of liability  
We accept no liability for any damage or malfunctions resulting from incorrect installation,  
inappropriate use of the device or failure to follow the instructions in this operating manual.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
3
Safety instructions  
Before you install the FMG90, read through this operating manual carefully. If the  
instructions contained within it are not followed, in particular the safety guidelines, this  
could result in danger for people, the environment, and the device and the system it is  
connected to.  
The FMG90 correspond to the state-of-the-art technology. This concerns the accuracy, the  
operating mode and the safe operation of the device.  
In order to guarantee that the device operates safely, the operator must act competently and  
be conscious of safety issues.  
Omega provides support for the use of its products either personally or via relevant literature.  
The customer verifies that our product is fit for purpose based on our technical information.  
The customer performs customer- and application-specific tests to ensure that the product is  
suitable for the intended use. With this verification all hazards and risks are transferred to our  
customers; our warranty is not valid.  
Qualified personnel:  
The personnel who are charged for the installation, operation and maintenance of the  
FMG90 must hold a relevant qualification. This can be based on training or relevant  
tuition.  
The personnel must be aware of this operating manual and have access to it at all times.  
The electrical connection should only be carried out by a fully qualified electrician.  
General safety instructions:  
In all work, the existing national regulations for accident prevention and safety in the  
workplace must be complied with. Any internal regulations of the operator must also be  
complied with, even if these are not mentioned in this manual.  
Degree of protection according to EN 60529:  
Please ensure that the ambient conditions at the site of use does not exceed the  
requirements for the stated protection rating (see Technical Data "10").  
Only use the FMG90 if it is in perfect condition. Damaged or faulty devices must be  
checked without delay and, if necessary, replaced.  
When fitting, connecting and removing the FMG90 use only suitable appropriate tools.  
Do not remove or obliterate type plates or other markings on the device, as otherwise the  
warranty is rendered null and void.  
Special safety instructions:  
(Further) Warnings that are specifically relevant to individual operating procedures or activities  
can be found at the beginning of the relevant sections of this operating manual.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
4
Construction and function  
Construction:  
The measuring tube with its earthing sleeves and electrodes  
passes through the sensor housing and forms the external  
process connection of the FMG90.  
A magnetic field for the measurement process is generated inside  
the sensor housing, which also contains the sensor and signal  
conditioning circuitry.  
The two stainless steel electrodes are located in the middle of the  
measuring tube between the earthing sleeves. The FMG90 does  
not need any moving parts to make measurements. The inside of  
the measuring tube is completely open, allowing the fluid to flow  
unhindered through the measuring tube.  
Function:  
The magnetic inductive flow sensor functions according to the induction principle:  
The measuring tube is located in a magnetic field (B). If  
an electrically conductive medium (V) flows through the  
measuring tube and, therefore, at right angles to the  
magnetic field, a voltage (U) which is proportional to  
the mean flow velocity will be induced in the medium  
and subsequently picked up by the two electrodes.  
5
Installation  
Before installing, check that  
ˆ the wetted materials of the device are suitable for the media being used (see Materials  
Table "10.1").  
ˆ the equipment is switched off and is in a safe and de-energised state.  
ˆ the equipment is depressurised and has cooled down.  
SUITABLE TOOLS:  
Use only suitable tools of the correct size.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
5.1 Installation instructions  
CAUTION!  
Risk of malfunction due to external magnetic fields!  
Magnetic fields close to the device can cause malfunctions and  
should be avoided.  
ª Ensure that no external magnetic fields are present at the  
installation site of the FMG90.  
The FMG90 can always be installed anywhere along the pipeline. Straight sections of  
piping are preferable, however.  
Installation can occur in horizontal and vertical pipes. The flow sensor is only suitable for  
application in completely filled pipe systems.  
As a matter of principle magnetic inductive flow sensors are widely independent from the  
flow profile. An inlet section is not absolutely necessary.  
To reach a most highly accuracy of the measurement, you should use straight inlet and  
outlet sections according to the nominal pipe diameter (DN). The inlet pipe section has to  
be at least 10 x DN; the outlet section 5 x DN in order to achieve the specified accuracy.  
The inlet and outlet sections and the gaskets must have the same or a slightly larger  
inside diameter than the measuring tube in order to achieve the specified accuracy.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
If two or more FMG90 devices are used side by side, maintain  
a separation of at least 2.5 cm between adjacent devices.  
If adjacent devices are too close together, operation of both  
devices may be impaired due to mutual interference.  
5.2 Assembly  
The FMG90 is installed directly into the pipeline. The compact design and light weight of the  
unit make wall-mounting unnecessary.  
IMPORTANT NOTICES:  
• Only use suitable gaskets for installation.  
• Observe the flow direction indicated on the type plate.  
• Observe the mounting dimensions.  
ª To ensure the best possible measuring accuracy, a horizontal  
installation position with increasing flow is preferable (no  
collecting of dirt deposits).  
ª Install the appropriate screwed connections at the installation  
location.  
ª Insert the FMG90 together with the gaskets.  
ª Screw the union nuts of the screwed connection onto the  
process connections of the FMG90.  
ª Tighten both union nuts with a maximum torque of 5 Nm.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
6 Electrical connection  
The electrical connection of the FMG90 is via the 4-pin plug M12x1 at the top.  
The corresponding connection cables with moulded coupling socket are available in various  
lengths.  
CAUTION! Electric current!  
The electrical connection should only be carried out by a fully qualified electrician.  
ª De-energize the electrical system before connecting the FMG90.  
Connection and wiring:  
ª Screw the coupling socket of the connection cable to the plug of the FMG90.  
ª Tighten the knurled nut of the coupling socket with a maximum torque of 1 Nm.  
ª Connect the connection cables according to the following wiring diagrams.  
Pin assignment:  
Pin 1: +UB  
Pin 3: GND  
Pin 2 / 4: Frequency output NPN/PNP  
Pin configuration with  
NPN frequency output:  
PNP frequency output:  
Pull-up- / pull-down-resistors R.  
We recommend using resistors of ~1 k(12V) respectively ~2,2 k(24V) and 0.25 W for  
the pull-up / pull-down wiring.  
Please note that the maximum signal current of 25 mA will not be exceeded.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
7 Commissioning and measuring operation  
Before switching on the FMG90 for the first time, please follow the instructions in the following  
section.  
7.1 Commissioning  
Check that  
ˆ the FMG90 has been installed correctly and that all screw connections are sealed.  
ˆ the electrical wiring has been connected properly.  
ˆ the measuring system is vented by flushing.  
ª Switch on the supply voltage.  
If the red LED lights permanently; then the FMG90 is ready for use and goes into measuring  
operation.  
7.2 Measuring operation  
The red LED is constantly lit to indicate that the FMG90 is  
operational.  
The green LED blinks according to the frequency of the output  
signal.  
This blinking is not perceptible to the human eye at frequencies  
above 30 to 40 Hz, so the green LED appears to be constantly lit.  
Frequency output:  
The frequency output provides a flow-  
proportional PNP/NPN square wave signal.  
8 Maintenance  
Maintenance:  
The FMG90 is maintenance-free and cannot be repaired by the user. In case of a defect, the  
device must be replaced or sent back the manufacturer for repair.  
CAUTION! Material damage!  
When opening the device, critical parts or components can be damage.  
ª Never open the device and perform any repair yourself.  
Cleaning:  
Clean the FMG90 with a dry or slightly damp lint-free cloth. Do not use sharp objects or  
aggressive agents for cleaning.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
9 Disassembly and disposal  
CAUTION! Risk of injury!  
Never remove the device from a plant in operation.  
ª Make sure that the plant is shut down professionally.  
Before disassembly:  
Prior to disassembly, ensure that  
ˆ the equipment is switched off and is in a safe and de-energised state.  
ˆ the equipment is depressurised and has cooled down.  
Disassembly:  
ª Remove the electrical connectors.  
ª Remove the FMG90 using suitable tools.  
Disposal:  
NO HOUSEHOLD WASTE!  
The FMG90 consists of various different materials. It must not be disposed of with  
household waste.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
10 Technical data  
The technical data of customised versions may differ from the data in these instructions.  
Please observe the information specified on the type plate.  
Type  
FMG 96  
Measurement device characteristics  
Flow range [GPM]  
3.3…66  
Accuracy  
Repeatability  
1% of reading  
1%  
Output signal  
starting from [GPM]  
1.3  
Max. flow rate [GPM]  
79.3  
Response time  
Indications  
< 100 ms  
red LED = Supply voltage • green LED = Flow  
Output signal characteristics  
Frequency output:  
- Pulse rate / K-  
Factor *  
15000  
3000  
1500  
750  
380  
300  
[pulses/gal]  
- Resolution *  
[ml/pulse]  
12.5  
- Signal shape  
Square wave signal • duty cycle 50:50  
can be connected as PNP or NPN open collector  
- Signal current  
25 mA  
Electrical characteristics  
Supply voltage  
Power consumption  
24 VDC ±15% or 12 VDC ±15%  
0.6 W  
Electrical protection  
measures  
short-circuit proof • protected against polarity reversal  
Electrical connection  
4-pin-plug M12x1  
Degree of protection  
IP 65 (only with a connected coupling)  
Process variables  
Medium to measure  
Water and other conductive liquids  
min. conductivity of  
the medium  
20 μS/cm  
Medium temp.  
Ambient temp.  
Nominal diameter  
Inner diameter  
14…140 °F (not freezing)  
41…140 °F  
DN 25  
0.98 inch  
Max. working  
145 psi (68 °F) • 116 psi (104 °F) • 87 psi (140 °F)  
pressure (at …)  
higher pressure ratings on demand  
Process connection  
½ - 14 NPT  
¾ - 14 NPT  
1 - 11.5 NPT  
1¼ - 11.5  
NPT  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
10.1 Materials table  
Component  
Material  
Component-  
wetted  
Sensor housing  
Measuring tube  
ABS  
X
Process connections  
X
X
X
X
O-ring  
Electrodes  
Grounding rings  
Stainless steel 316L  
10.2 Pressure drop  
FMG 91 / FMG 92  
FMG 93  
FMG 96  
FMG 94 / FMG 95  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
10.3 Dimensions  
FMG 91 - 95  
L1  
L2  
D1  
D2  
Type  
FMG 91  
FMG 92  
FMG 93  
4± 0.02  
4± 0.02  
0.83± 0.01 ½ - 14 NPT  
0.83± 0.01 ½ - 14 NPT  
Ø 0.31  
Ø 0.31  
Ø 0.55  
Ø 0.71  
Ø 0.71  
4.02± 0.02 0.83± 0.01 ¾ - 14 NPT  
1 – 11.5  
4.41± 0.02 0.98± 0.01 NPT  
1 – 11.5  
FMG 94  
FMG 95  
4.41± 0.02 0.98± 0.01 NPT  
FMG 96  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Download from Www.Somanuals.com. All Manuals Search And Download.  
M-5204/1112  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Milwaukee Sander HEAVY DUTYSANDERS User Manual
Minicom Advanced Systems Switch 5UM70166 User Manual
Minolta Universal Remote SRU 3040 10 User Manual
Nokia Bluetooth Headset LPS 5 User Manual
Numark Industries DJ Equipment DM2000X User Manual
NuTone Door LA 156PR User Manual
Oki Network Card 8100e User Manual
Omega Vehicle Security Heat Pump PX791 User Manual
Onkyo Stereo System TX DS575x User Manual
Panasonic Switch FP X Series User Manual