NI PXI/PCI-5114 Specifications
8-Bit 250 MS/s Digitizer
Unless otherwise noted, the following conditions were used for each
specification:
•
•
•
All filter settings
All impedance selections
Sample clock set to 250 MS/s
Typical values are representative of an average unit operating at room
temperature. Specifications are subject to change without notice. For the
most recent NI 5114 specifications, visit ni.com/manuals.
To access the NI 5114 documentation, including the NI High-Speed
Digitizers Getting Started Guide, which contains functional descriptions of
the NI 5114 signals, navigate to Start»Programs»National Instruments»
NI-SCOPE»Documentation.
Phase-Locked Loop (PLL) Reference Clock................................... 7
TRIG (External Trigger, Front Panel Connector)............................ 10
PFI 0 and PFI 1 (Programmable Function Interface,
AUX Front Panel Connectors)............................................................. 10
TClk Specifications................................................................................. 12
Waveform Specifications........................................................................ 13
Calibration............................................................................................... 14
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
Voltage Levels
Full Scale (FS)
InputRangeand
Programmable
Vertical Offset
50 Ω
Vertical Offset
1 MΩ
Vertical Offset
—
Range
Range
(Vpk-pk
0.04
0.1
0.2
0.4
1
)
Range (V)
(Vpk-pk
0.04
0.1
0.2
0.4
1.0
2.0
4.0
10
)
Range (V)
0.8
0.8
0.8
0.8
0.8
0.8
0.8
0.8
6.5
8.0
2
6.0
8.0
4
5.0
8.0
30
10
2.0
20
25
40
15
MaximumInput
Overload
50 Ω
1 MΩ
|Peaks| ≤ 35 V
—
—
7 Vrms with |Peaks| ≤ 10 V
Accuracy
Resolution
8 bits
DC Accuracy
(Programmable
Vertical
NI PXI-5114:
(1.5% of Input + 0.3% of FS + 200 μV)
Within 5 °C of
self-calibration
temperature.
NI PCI-5114:
(1.5% of Input + 0.3% of FS + 280 μV)
Offset = 0 V)
Programmable
Vertical Offset
Accuracy
2% of offset setting
Within 5 °C of
self-calibration
temperature.
DC Drift
(0.03% of Input + 0.06% of FS + 40 μV) per °C
≤ –60 dB at 10 MHz
—
Crosstalk,
Typical
CH 0 to/from
CH 1, External
Trigger to CH 0
or CH 1.
≤ –45 dB at 100 MHz
© National Instruments Corporation
3
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
Bandwidth and Transient Response
Bandwidth
(–3 dB)
Range
Rise/FallTime,
Typical
—
(Vpk-pk
)
Bandwidth
125 MHz
All ranges
2.8 ns
except 0.04
0.04
100 MHz
3.5 ns
Bandwidth
Limit Filter
—
20 MHz Noise Filter
AC Coupling*
Cutoff (–3 dB),
Typical
*AC coupling
available on
1 MΩ only.
12 Hz
Passband
Flatness
Referenced to
50 kHz.
1 dB up to 50 MHz
Bandwidth
limit filter off.
1
0
–1
–2
–3
–4
–5
–6
–7
–8
–9
Full Bandwidth
Noise Filter
–10
1 M
10 M
100 M
200 M
Frequency (Hz)
Figure 1. NI 5114 Frequency Response (Typical)
NI PXI/PCI-5114 Specifications
4
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
Spectral Characteristics
Range (Vpk-pk
)
10 MHz,
–1 dBFS
input signal.
Spurious Free
All ranges except 0.04
0.04
Dynamic Range
with Harmonics
(SFDR), Typical
Includes the
2nd through
the 5th
–58 dBc
–58 dBc
–58 dBc
Total Harmonic
Distortion
(THD), Typical
harmonics.
–58 dBc
6.2
Measured
from DC to
125 MHz.
Effective
Number of Bits
(ENOB),
7.2
20 MHz
bandwidth
limit filter off.
Calculated*
Signal to Noise
and Distortion
(SINAD),
44 dB
38 dB
Typical
RMS Noise
Range
(Vpk-pk
20 MHz
Filter On
20 MHz
Filter Off
50 Ω
)
terminator
connected
to input.
All ranges
except 0.04
0.28% FS
0.28% FS
0.04
0.28% FS
0.45% FS
*ENOB = log2(sinad) – ½log2(1.5) – log2(A/V)
where
sinad = the linear representation of SINAD
A = amplitude of the supplied sine wave during the test
V = (peak) full-scale range of the waveform recorder input
Refer to 1057-1994 IEEE Standard for Digitizing Waveform Recorders for information on equation derivation.
© National Instruments Corporation
5
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Horizontal
Sample Clock
Specification
Value
Comments
Sources
Internal, Onboard Clock (internal VCXO)*
External, CLK IN (front panel SMB connector)
* Internal Sample
Clock is locked to the
Reference Clock or
derived from the
onboard VCXO.
Onboard Clock (Internal VCXO)
Sample Rate
Range
Real-Time Sampling
(Single Shot)
Random Interleaved
Sampling (RIS)
* Divide by n
decimation used
for all rates less than
250 MS/s. For more
information about
Sample Clock and
decimation, refer to
the NI High-Speed
Digitizers Help.
3.815 kS/s to 250MS/s* 250 MS/s to 5 GS/s in
increments of 250 MS/s
Timebase
Frequency
250 MHz
When not using
External Sample
Clock.
Timebase
Accuracy
Not Phase-Locked to
Reference Clock
Phase-Locked to
Reference Clock
—
25 ppm
Equal to the Reference
Clock accuracy
Sample Clock
Delay Range
1 Sample Clock period
≤ 20 ps
—
—
Sample Clock
Delay Resolution
NI PXI/PCI-5114 Specifications
6
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
External Sample Clock
Sources
CLK IN (front panel SMB connector)
50 MHz to 250 MHz
—
Frequency
Range
Divide by n
decimation available
where
1 ≤ n ≤ 65,535.
Formoreinformation
about Sample Clock
and decimation, refer
to the NI High-Speed
Digitizers Help.
Duty Cycle
Tolerance
45% to 55%
—
Phase-Locked Loop (PLL) Reference Clock
Specification
Sources
Value
Comments
NI PXI-5114
NI PCI-5114
—
PXI_CLK10 (backplane
connector)
RTSI 7
CLK IN (front panel SMB
CLK IN (front panel SMB connector)
connector)
Frequency
Range
1 MHz to 20 MHz in 1 MHz increments.
Default of 10 MHz.
—
The PLL Reference Clock frequency must be accurate
to 50 ppm.
Duty Cycle
Tolerance
45% to 55%
—
—
Exported
NI PXI-5114
NI PCI-5114
Reference Clock
Destinations
PFI <0..1> (front panel
9-pin mini-circular
DIN connector)
PFI <0..1> (front panel
9-pin mini-circular
DIN connector)
PXI_Trig <0..7>
RTSI <0..7>
(backplane connector)
© National Instruments Corporation
7
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
CLK IN (Sample Clock and Reference Clock Input,
Front Panel Connector)
Specification
Value
Comments
Input Voltage
Range
Sine wave: 0.65 Vpk-pk to 2.8 Vpk-pk (0 dBm to 13 dBm)
Square wave: 0.2 Vpk-pk to 2.8 Vpk-pk
—
Maximum Input
Overload
7 Vrms with |Peaks| ≤ 10 V
—
Impedance
Coupling
50 Ω
—
—
AC
Trigger
Reference (Stop) Trigger
Specification
Value
Comments
Trigger Types
and Sources
Types
Sources
Refer to the
following
sections and to
NI High-Speed
Digitizers Help
for more
Edge, Window, Hysteresis,
Video, Digital, Immediate,
and Software
CH 0, CH 1, TRIG,
PXI_Trig<0..6>, PFI <0..1>,
PXI Star Trigger,
RTSI<0..6>, and Software
information.
Time
Resolution
TDC
On
Onboard Clock
External Clock
N/A
TDC = Time to
Digital
Conversion
Circuit.
40 ps
4 ns
Off
External Clock Period
Rearm Time
10 μs
Rearm Time
TDC
On
Holdoff set to 0
Off
2 μs
Holdoff
From Rearm Time up to [(235 – 1) × (Sample Clock Period)]
—
—
Trigger Delay
From 0 up to [(235 – 1) – posttrigger samples] ×
(1/ sample rate), in seconds
NI PXI/PCI-5114 Specifications
8
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
Analog Trigger (Edge, Window, and Hysteresis Trigger Types)
Sources
CH 0 (front panel BNC connector)
CH 1 (front panel BNC connector)
TRIG (front panel BNC connector)
8 bits (1 in 256)
—
Trigger Level
Resolution
CH 0, CH 1
100% FS
TRIG (External Trigger)
—
Trigger Level
Range
5 V
Edge Trigger
Sensitivity
5% FS up to 100 MHz
5% FS up to 10 MHz
0.5 Vpk-pk up to 100 MHz
0.5 V up to 10 MHz
Level
Accuracy,
Typical
Jitter
≤ 65 ps rms
—
—
Trigger Filters Low Frequency (LF) Reject High Frequency (HF) Reject
50 kHz
50 kHz
Digital Trigger (Digital Trigger Type)
Sources
NI PXI-5114
NI PCI-5114
—
PXI_Trig <0..6> (backplane RTSI <0..6>
connector)
PFI <0..1> (front panel
PFI <0..1> (front panel
SMB connector)
SMB connector)
PXI Star Trigger (backplane
connector)
Video Trigger (Video Trigger Type)
Sources CH 0 (front panel BNC connector)
CH 1 (front panel BNC connector)
TRIG (front panel BNC connector)
—
© National Instruments Corporation
9
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
Value
Comments
Types
Specific Line
Any Line
—
Specific Field
Standards
SDTV: M-NTSC, B/G-PAL, SECAM, M-PAL
fps = fields per
second;
EDTV: 480i/59.94 fps, 480i/60 fps, 480p/59.94 Fps,
480p/60 Fps, 576i/50 fps, 576p/50 Fps
Fps = Frames
per second
HDTV: 720p/50 Fps, 720p/59.94 Fps, 720p/60 Fps,
1080i/50 fps, 1080i/59.94 fps, 1080i/60 fps, 1080p/24 Fps
TRIG (External Trigger, Front Panel Connector)
Specification
Connector
Value
Comments
BNC
—
—
—
—
Impedance
Coupling
1 MΩ in parallel with 22 pF
AC, DC
12 Hz
AC-Coupling Cutoff
(–3 dB)
Input Voltage Range
5 V
—
—
Maximum Input Overload |Peaks| ≤ 42 V
PFI 0 and PFI 1 (Programmable Function Interface,
AUX Front Panel Connectors)
Specification
Connector
Value
Comments
9-pin mini-circular DIN
Bi-directional
—
—
Direction
NI PXI/PCI-5114 Specifications
10
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Specification
As an Input (Trigger)
Destinations
Value
Comments
Start Trigger (Acquisition Arm)
Reference (Stop) Trigger
Arm Reference
Advance Trigger
150 kΩ
—
Input Impedance
—
—
—
—
VIH
VIL
2.0 V
0.8 V
Maximum Input
Overload
–0.5 V to 5.5 V
Maximum
Frequency
25 MHz
—
—
As an Output (Event)
Sources
Start Trigger (Acquisition Arm)
Reference (Stop) Trigger
End of Record
Done (End of Acquisition)
Probe Compensation (1 kHz, 50% duty cycle
square wave, PFI 1 only)
Output Impedance
Logic Type
50 Ω
—
—
—
3.3 V CMOS
24 mA
Maximum Drive
Current
Maximum
Frequency
25 MHz
—
© National Instruments Corporation
11
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
TClk Specifications
National Instruments TClk synchronization method and the NI-TClk driver
are used to align the sample clocks on any number of SMC-based modules
in a chassis. For more information about TClk synchronization, refer to the
NI-TClk Synchronization Help, which is located within the NI High-Speed
Digitizers Help.
•
Specifications are valid for any number of modules installed in one
NI PXI-1042 chassis.
•
•
•
All parameters set to identical values for each SMC-based module.
Sample Clock set to 250 MS/s and all filters are disabled.
For other configurations, including multichassis systems, contact
NI Technical Support at ni.com/support.
Note Although you can use NI-TClk to synchronize nonidentical modules, these
specifications apply only to synchronizing identical modules.
Specification
Value
Comments
Intermodule SMC Synchronization Using NI-TClk for Identical Modules (Typical)
Skew
500 ps
Caused by clock and analog
path delay differences.
No manual adjustment
performed.
Skew After Manual < 20 ps
Adjustment
For information about
manual adjustment, refer
to the Synchronization
Repeatability Optimization
topic in the NI-TClk
Synchronization Help.
For additional help with the
adjustment process, contact
NI Technical Support at
ni.com/support.
Sample Clock
Adjustment
Resolution
< 20 ps
—
NI PXI/PCI-5114 Specifications
12
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Waveform Specifications
Specification
Value
8 megasamples
Comments
Onboard Memory
Size
8 MB per Channel
Standard
—
per channel
64 MB per Channel
Option
64 megasamples
per channel
256 MB per Channel
Option
256 megasamples
per channel
Minimum Record
Length
1 Sample
—
Number of
Pretrigger Samples
Zero up to full Record Length
Zero up to full Record Length
Single-record
mode and
multiple-record
mode.
Number of
Posttrigger Samples
Single-record
mode and
multiple-record
mode.
Maximum Number
of Records in
Onboard Memory
8 MB/channel
32 MB/channel
256 MB/channel
32,768
—
262,144
1,048,576
Allocated Onboard
Memory per Record
(Record Length × 1 byte/S) + 240 bytes,
rounded up to next multiple of 128 bytes
—
or
256 bytes, whichever is greater
© National Instruments Corporation
13
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Calibration
Specification
Value
Comments
Self-Calibration
Self-calibration is done on software command.
The calibration corrects for gain, offset,
—
compensated 1 MΩ attenuator, triggering, and
timing adjustment errors for all input ranges.
External Calibration
(Factory Calibration)
The external calibration calibrates the VCXO, gain,
and the voltage reference. Appropriate constants are
stored in nonvolatile memory.
—
Interval for External
Calibration
2 years
—
—
Warm-Up Time
15 minutes
Power
Specification
Typical Value
Comments
NI PXI-5114
840 mA
1.1 A
NI PCI-5114
1.6 A
—
+3.3 VDC
+5 VDC
1.7 A
+12 VDC
–12 VDC
Total Power
250 mA
170 mA
13.32 W
45 mA
—
14.32 W
NI PXI/PCI-5114 Specifications
14
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Software
Specification
Value
Comments
Driver Software
NI PXI-5114: NI-SCOPE 2.9 or later.
NI PCI-5114: NI-SCOPE 3.1 or later.
—
NI-SCOPE is an IVI-compliant driver that allows you to
configure, control, and calibrate the NI 5114. NI-SCOPE
provides application programming interfaces for many
development environments.
Application
Software
NI-SCOPE provides programming interfaces,
documentation, and examples for the following
application development environments:
—
•
•
•
•
•
LabVIEW
LabWindows™/CVI™
Measurement Studio
Microsoft Visual C/C++
Microsoft Visual Basic
Interactive Soft
Front Panel and
Configuration
The Scope Soft Front Panel 2.3 or later supports interactive
control of the NI 5114. The Scope Soft Front Panel is
included on the NI-SCOPE CD.
—
National Instruments Measurement & Automation Explorer
(MAX) also provides interactive configuration and test
tools for the NI 5114. MAX is included on the NI-SCOPE
CD.
© National Instruments Corporation
15
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Environment
NI PXI-5114
Note To ensure that the NI PXI-5114 cools effectively, follow the guidelines in
the Maintain Forced Air Cooling Note to Users included in the NI PXI-5114 kit.
The NI PXI-5114 is intended for indoor use only.
Specification
Value
Comments
Operating
0 ºC to +55 ºC in all NI PXI chassis except the following:
—
Temperature
0 ºC to +45 ºC when installed in an NI PXI-1000/B or
PXI-101x chassis.
Meets IEC-60068-2-1 and IEC-60068-2-2.
Storage
Temperature
–40 ºC to +71 ºC. Meets IEC-60068-2-1 and
IEC-60068-2-2.
—
—
—
—
—
Operating Relative
Humidity
10% to 90%, noncondensing. Meets IEC-60068-2-56.
Storage Relative
Humidity
5% to 95%, noncondensing. Meets IEC-60068-2-56.
Operating Shock
30 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.
Storage Shock
50 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.
Operating Vibration 5 Hz to 500 Hz, 0.31 grms. Meets IEC-60068-2-64.
—
—
Storage Vibration
5 Hz to 500 Hz, 2.46 grms. Meets IEC-60068-2-64.
Test profile exceeds requirements of MIL-PRF-28800F,
Class B.
Altitude
2,000 m maximum (at 25 °C ambient temperature)
2
—
—
Pollution Degree
NI PXI/PCI-5114 Specifications
16
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
NI PCI-5114
Note To ensure that the NI PCI-5114 cools effectively, follow the guidelines in the
Maintain Forced Air Cooling Note to Users included in the NI PCI-5114 kit. Also, to
maximize airflow and extend the life of the device, leave any adjacent PCI slots empty.
The NI PCI-5114 is intended for indoor use only.
Specification
Operating
Value
Comments
0 ºC to +45 ºC. Meets IEC-60068-2-1 and IEC-60068-2-2.
—
Temperature
Storage
Temperature
–40 ºC to +71 ºC. Meets IEC-60068-2-1 and
IEC-60068-2-2.
—
—
—
—
—
Operating Relative
Humidity
10% to 90%, noncondensing. Meets IEC-60068-2-56.
Storage Relative
Humidity
5% to 95%, noncondensing. Meets IEC-60068-2-56.
Storage Shock
50 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.
Storage Vibration
5 Hz to 500 Hz, 2.46 grms. Meets IEC-60068-2-64.
Test profile exceeds requirements of MIL-PRF-28800F,
Class B.
Altitude
2,000 m maximum (at 25 °C ambient temperature).
2
—
—
Pollution Degree
© National Instruments Corporation
17
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Safety, Electromagnetic Compatibility,
and CE Compliance
Specification
Value
Comments
Safety
The NI 5114 meets the requirements of the following
standards of safety for electrical equipment for measurement,
control, and laboratory use:
—
•
•
IEC 61010-1, EN 61010-1
UL 61010-1, CAN/CSA-C22.2 No. 61010-1
For UL and other safety certifications, refer to the product label or visit ni.com/certification,
search by model number or product line, and click the appropriate link in the Certification column.
Emissions
EN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
—
Immunity
EMC/EMI
EN 61326 EMC requirements; Minimum Immunity
CE, C-Tick, and FCC Part 15 (Class A) Compliant
—
—
Note: For EMC compliance, you must operate this device
with shielded cabling.
This product meets the essential requirements of applicable European Directives, as amended
for CE marking, as follows:
Low-Voltage
Directive
(safety)
73/23/EEC
—
Electro-
89/336/EEC
—
magnetic
Compatibility
Directive
(EMC)
For full EMC compliance, you must operate this device with shielded cabling. Refer to the
Declaration of Conformity (DoC) for this product for any additional regulatory compliance
information. To obtain the DoC for this product, visit ni.com/certification, search by model
number or product line, and click the appropriate link in the Certification column.
Waste
EU Customers: At the end of their life cycle, all products must be sent to a
WEEE recycling center. For more information about WEEE recycling centers
and National Instruments WEEE initiatives, visit ni.com/environment/
weee.htm.
Electrical and
Electronic
Equipment
(WEEE)
NI PXI/PCI-5114 Specifications
18
ni.com
Download from Www.Somanuals.com. All Manuals Search And Download.
Physical
Specification
Value
Comments
Dimensions
NI PXI-5114
3U, One slot,
PXI/cPCI Module
21.6 × 2.0 × 13.0 cm
(8.5 × 0.8 × 5.1 in.)
NI PCI-5114
—
35.5 × 2.0 × 11.3 cm
(13.4 × 0.8 × 4.4 in.)
Weight
455 g (16 oz)
421 g (14.8 oz)
—
—
Front Panel Connectors
Label
Function
Analog Input
Analog Input
Connector Type
BNC female
CH 0
CH 1
BNC female
TRIG
CLK IN
External Trigger
BNC female
Sample Clock Input and
Reference Clock Input
SMB jack
AUX I/O
PFI 0, PFI 1
9-pin mini-circular DIN
NI PXI-5114 Front Panel Indicators
Label
Function
For more
information,
refer to the
NI High-Speed
Digitizers Help.
ACCESS
The ACCESS LED indicates the status of the PCI bus and
the interface from the NI PXI-5114 to the controller.
ACTIVE
The ACTIVE LED indicates the status of the onboard
acquisition hardware of the NI PXI-5114.
© National Instruments Corporation
19
NI PXI/PCI-5114 Specifications
Download from Www.Somanuals.com. All Manuals Search And Download.
Where to Go for Support
The National Instruments Web site is your complete resource for technical
support. At ni.com/support you have access to everything from
troubleshooting and application development self-help resources to email
and phone assistance from NI Application Engineers.
A Declaration of Conformity (DoC) is our claim of compliance with the
Council of the European Communities using the manufacturer’s
declaration of conformity. This system affords the user protection for
electronic compatibility (EMC) and product safety. You can obtain the DoC
for your product by visiting ni.com/certification. If your product
supports calibration, you can obtain the calibration certificate for your
product at ni.com/calibration.
National Instruments corporate headquarters is located at
11500 North Mopac Expressway, Austin, Texas, 78759-3504.
National Instruments also has offices located around the world to help
address your support needs. For telephone support in the United States,
create your service request at ni.com/support and follow the calling
instructions or dial 512 795 8248. For telephone support outside the United
States, contact your local branch office:
Australia 1800 300 800, Austria 43 0 662 45 79 90 0,
Belgium 32 0 2 757 00 20, Brazil 55 11 3262 3599,
Canada 800 433 3488, China 86 21 6555 7838,
Czech Republic 420 224 235 774, Denmark 45 45 76 26 00,
Finland 385 0 9 725 725 11, France 33 0 1 48 14 24 24,
Germany 49 0 89 741 31 30, India 91 80 41190000,
Israel 972 0 3 6393737, Italy 39 02 413091, Japan 81 3 5472 2970,
Korea 82 02 3451 3400, Lebanon 961 0 1 33 28 28,
Malaysia 1800 887710, Mexico 01 800 010 0793,
Netherlands 31 0 348 433 466, New Zealand 0800 553 322,
Norway 47 0 66 90 76 60, Poland 48 22 3390150,
Portugal 351 210 311 210, Russia 7 095 783 68 51,
Singapore 1800 226 5886, Slovenia 386 3 425 4200,
South Africa 27 0 11 805 8197, Spain 34 91 640 0085,
Sweden 46 0 8 587 895 00, Switzerland 41 56 200 51 51,
Taiwan 886 02 2377 2222, Thailand 662 278 6777,
United Kingdom 44 0 1635 523545
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation.
Refer to the Terms of Use section on ni.com/legal for more information about National
Instruments trademarks. Other product and company names mentioned herein are trademarks or trade
names of their respective companies. For patents covering National Instruments products, refer to the
appropriate location: Help»Patents in your software, the patents.txt file on your CD, or
ni.com/patents.
© 2005–2006 National Instruments Corporation. All rights reserved.