Leica DM4000 B
Leica DM4000 M
Leica DM5000 B
Operating Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Leica DM4000 B
Leica DM4000 M
Leica DM5000 B
Operating Manual
Download from Www.Somanuals.com. All Manuals Search And Download.
Copyrights
Copyrights
All rights to this documentation are held by Leica
Microsystems Wetzlar GmbH. Reproduction of
text or illustrations (in whole or in part) by print,
photocopy, microfilm or other methods (includ-
ing electronic systems) is not allowed without
express written permission from Leica
Microsystems Wetzlar GmbH.
The term "Windows" can be used in the following
text without further identification. It is
a
registered trademark of the Microsoft
Corporation. Otherwise, no inference with
regard to the free usability of product names
may be drawn from the use of those names.
The instructions contained in the following docu-
mentation reflect state-of-the-art techno-logy
and knowledge standards. We have compiled
the texts and illustrations as accurately as
possible. Nevertheless, no liability of any kind
may be assumed for the accuracy of this manu-
al’s contents. Still, we are always grateful for
comments and suggestions regarding potential
mistakes within this documentation.
The information in this manual is subject to modifi-
cation at any time and without notification.
Download from Www.Somanuals.com. All Manuals Search And Download.
Contents
Contents
6. Startup ........................................................ 31
6.1 Functional Principle.................................. 31
6.2 Switching on the Microscope ................ 34
6.3 The Display
(Leica DM4000 B/DM4000 M) ................. 35
6.4 The Function Keys .................................... 36
6.5 Köhler Illumination .................................... 37
6.6. Checking Phase Contrast Rings ............. 39
6.7 Adjusting the Light Sources .................... 40
1.
Important Notes about this Manual .....
7
2. Safety Notes ..............................................
2.1. General Safety Notes ...............................
2.2. Electrical Safety ........................................
8
8
8
3. Overview of the Instrument .................... 10
4. Unpacking the Microscope .................... 14
7. Operation ................................................... 46
7.1 Switching on the Microscope ................ 46
7.2 Stages and Specimen Displacement .... 46
7.3 Focusing ..................................................... 47
7.4 Tubes...........................................................
48
5. Assembling the Microscope .................. 16
5.1 Stage ........................................................... 17
5.2 Condenser .................................................. 18
5.3 Tube and Eyepieces ................................. 19
5.4 Objectives .................................................. 19
5.5 Light Sources for the
7.5 Eyepieces ................................................... 49
7.6 Objectives .................................................. 50
7.7 Magnification Changer ............................ 51
7.8 Light Sources ............................................. 52
7.9 Aperture Diaphragm and
Transmitted Light Axis ............................. 20
5.6 Light Sources for the
Incident Light Axis .................................... 21
5.7 Equipping the
Incident Light filter turret ........................ 26
5.8 Polarizer and Analyzer ............................. 27
5.9 DIC Prisms .................................................. 28
5.10 Optional Accessories ............................... 29
5.11 Connection to the Power Supply............ 30
5.12 Connection to the
Field Diaphragm ........................................ 52
8. Imaging Procedure for
Leica DM4000 B/Leica DM5000 B ......... 53
8.1 Transmitted Light ...................................... 53
8.1.1 Bright Field ...................................... 53
8.1.2 Phase Contrast ............................... 53
8.1.3 Dark Field ......................................... 54
8.1.4 Polarization ..................................... 55
8.1.5 Differential
CTR5000 Electronics Box ......................... 30
Interference Contrast .................... 56
5
Download from Www.Somanuals.com. All Manuals Search And Download.
Contents
8.2 Fluorescence ............................................. 57
9. Imaging Procedure for
Leica DM4000 M ....................................... 58
9.1 Incident Light ............................................. 58
12. Essential
Wear and Spare Parts ............................. 66
9.1.1 Bright Field ...................................... 58 13. Abbreviations and Pictograms .............. 67
9.1.2 Dark Field......................................... 58
9.1.3 Polarization ..................................... 59 14. Index ........................................................... 68
9.1.4 Interference Contrast .................... 60
9.2 Transmitted Light ...................................... 60 15. EU Declaration of Conformity ................ 69
9.2.1 Bright Field ...................................... 60
10. Trouble Shooting ...................................... 61
11. Care of the Microscope ........................... 64
11.1 Dust Cover .................................................. 64
11.2 Cleaning ...................................................... 64
11.3 Handling Acids and Bases ...................... 65
6
Download from Www.Somanuals.com. All Manuals Search And Download.
1. Important Notes about this Manual
1. Important Notes about this Manual
Caution!
This operating manual contains important in-
This operating manual is an essential com-
structions and information for the operational
ponent of the microscope, and must be read
safety and maintenance of the microscope and
carefully before the microscope is put into
accessories. Therefore, it must be kept and
operation or used.
taken care of.
Text symbols and their meanings:
(1.2)
Numbers in parentheses, such as "(1.2)", corre-
spond to illustrations (in the example, Figure 1,
Item 2).
→ p. 20
Numbers with pointer arrows (for example
→ p.20), point to a certain page of this manual.
Special safety instructions are indicated
with the triangle symbol shown here, and
have a gray background.
Caution! The microscope and accessories can
be damaged when operated incorrectly.
!
Explanatory note.
Item not contained in all configurations.
*
7
Download from Www.Somanuals.com. All Manuals Search And Download.
2. Safety Notes
2. Safety Notes
2.2 Electrical Safety
2.1 General Safety Notes
General specifications
This safety class 1 device is constructed and
tested in accordance with EN 61010-1/IEC 1010-1,
safety regulations for electrical measuring, con-
trol, and laboratory devices.
Leica CTR5000 electronics box (for DM5000 B)
For indoor use only.
Supply voltage:
Frequency:
Power input:
Fuses:
90-250 V~
50-60 Hz
max. 290 VA
T6,3 A
Caution!
(IEC 60127-2/3)
15-35°C
max. 80% to 30°C
II
In order to maintain this condition and to en-
sure safe operation, the user must follow the
instructions and warnings contained in this
operating manual.
Ambient temperature:
Relative humidity:
Overvoltage category:
Pollution degree:
2
Microscope
Caution!
For indoor use only.
Supply voltage:
Frequency:
Power input:
DM4000
DM5000
Fuses:
DM4000
The devices and accessories described in
this operating manual have been tested for
safety and potential hazards.
The responsible Leica affiliate or the main
plant in Wetzlar must be consulted when-
ever the device is altered, modified or used
in conjunction with non-Leica components
that are outside of the scope of this manual.
90-250 V~
50-60 Hz
max. 180 VA
max. 290 VA
T6,3 A
(IEC 60127-2/3)
DM5000
See CTR5000
Unauthorized alterations to the device or
noncompliant use shall void all rights to any
warranty claims!
Ambient temperature:
Relative humidity:
Overvoltage category:
Pollution degree:
15-35°C
max. 80% to 30°C
II
2
8
Download from Www.Somanuals.com. All Manuals Search And Download.
2. Safety Notes
Supply unit ebq 100
Caution!
For indoor use only.
Supply voltage:
Frequency:
Power input:
Fuses:
Ambient temperature:
Relative humidity:
Overvoltage category:
Pollution degree:
(see enclosed manual)
90-250 V~
50-60 Hz
max. 155 VA
2xT2A (IEC 127)
15-35°C
max. 80% to 30°C
II
2
Never use any fuses as replacements other
than those of the types and the current rat-
ings listed here. Using patched fuses or
bridging the fuse holder is not permitted.
Caution!
The microscope’s electrical accessory com-
ponents are not protected against water.
Water can cause electric shock.
Caution!
Caution!
The power plug may only be plugged into an
outlet equipped with a grounding contact.
Protect the microscope from excessive tem-
perature fluctuations. Such fluctuations can
lead to the accumulation of condensation,
which can damage the electrical and optical
components.
Do not interfere with the grounding function
by using an extension cord without a ground
wire. Any interruption of the ground wire in-
side or outside of the device, or release of
the ground wire connection, can cause the
device to become hazardous. Intentional
ground interruption is not permitted!
Ambient temperature: 15-35°C.
Caution!
Before exchanging the fuses or lamps, be
absolutely certain to switch-off the main
power switch and remove the power cable.
Caution!
Through connection to the grounding con-
nection, ancillary equipment with its own
and/or extra power supply may be brought to
the same ground wire potential. For
connections without a ground connector,
Leica Service must be consulted.
9
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Overview of the Instrument
3. Overview of the Instrument
Leica DM4000 B / DM5000 B
Specification
Leica DM4000 M
• transmitted light: BF, DF, PH,
Pol (DM5000 B also ICT)
Imaging Procedure
• transmitted BF, DF, PH
light:
ICT, Pol
• incident light: fluorescence
• incident light: BF, DF, ICR, Pol
Transmitted Light Axis
Incident Light Axis
• automatic Illumination Manager
(motorized aperture diaphragm and field diaphragm,
motorized intensity control)
• automatic Constant Color Intensity Control (CCIC)
• motorized shutter
• integrated into the stand
• motorized 5x filter turret
(DM5000 B 8x optional)
• integrated into the stand
• motorized 4x filter turret
• automatic Illumination
Manager
• with
FIM
(Fluorescence
Intensity Managemer) for de-
creasing light intensity in 5
stages
• motorized shutter
• mechanical “Booster Lens”
for increasing fluorescence
intensity
• motorized shutter
Z Pinion
• manual
• manual
Objective nosepiece
• manual
• absolute coded
• absolute encoded
• 6x with M32 thread
• slot for DIC prisms
and Pol compensators
(optional)
• 6x with M25 thread
(DM5000 B: 7x; mot. DIC
objective prism turret with 4
positions optional)
• manual
• replaceable specimen stage
• coaxial pinion length: 155 mm
X/Y Stage
Tube
• manual
• replaceable specimen stage
• coaxial pinion length: 140 mm
• manual or motorized
• optionally with two camera outputs
10
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Overview of the Instrument
Specification
Condenser
Leica DM4000 B / DM5000 B
Leica DM4000 M
• motorized condenser head
• motorized condenser turret for light rings,
DF stop, DIC prisms
• optional polarizer integrated and motorized
• automatic Köhler Illumination
Magnification Changer
Control Panels
• manual
• absolute coded
• 1x; 1.25x; 1.6x
• manual
• absolute coded
• 1x; 1.5x; 2x
• operating buttons for all motorized microscope functions
• additional variable function keys
• focusing knobs
• LC display
• DM5000 B with LeicaScreen (touchscreen)
Computer Interface
Software Tools
• RS232C
• Leica DMControl for WindowsTM 2000, XP, NT;
• with plugins for:
• customitsation
• DM Operation
(remote control)
• basic Image Viewer
CTR5000
For Leica DM5000 B only:
Electronics Box
Separate control unit with
power supply for 100W halogen
lamp
see p. 8 (electrical safety)
11
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Overview of the Instrument
1
2
3
14
4
5
6
7
13
12 11 10 9 8
Fig. 1 Leica DM4000 M left side of the stand with AET22 advanced ergotube
1
2
3
4
5
6
7
Eyepiece
8
9
Function keys field diaphragm
Eyepiece tube
Transmitted light/incident light switch
Tube
10 Function keys aperture diaphragm
11 Function keys: Light intensity
Objective nosepiece with objectives
Specimen stage with specimen holder
Condenser
12 Focus dial with coarse and fine adjustment
13 Variable function keys (factory pre-assigned)
14 Lamp adjustment window
LC display
12
Download from Www.Somanuals.com. All Manuals Search And Download.
3. Overview of the Instrument
15
22
16
21 20
19 18
17
Fig. 2
Leica DM4000 B right side of the stand with Advanced Ergotube AET22
15 Lamp housing for incident light
16 Lamp housing for transmitted light
17 Transmitted light filter, optional
18 Transmitted light filter, optional
19 Variable function keys (factory pre-assigned)
20 X/Y coaxial drive, height adjustable
21 Focus fine adjustment
22 Motorized filter cube exchanger
13
Download from Www.Somanuals.com. All Manuals Search And Download.
4. Unpacking the Microscope
4. Unpacking the Microscope
The external ebq 100 supply unit* is delivered in
The device is delivered in two boxes.
separate packaging.
The stand box contains the following compo-
For the Leica DM5000 B microscope:
The CTR5000 electronics box is also delivered in
separate packaging.
nents:
• Stand with integrated incident light axis and
objective nosepiece
First, carefully remove all components from the
transportation and packaging materials.
• Specimen stage with stage bracket
• Power cable and PC connecting cable
• CD with Leica software package
Note:
Avoid touching the lens surfaces of the
objectives. If fingerprints do appear on the glass
surfaces, remove them with a soft leather or
linen cloth. Even small traces of finger
perspiration can damage the surfaces of optical
surfaces in a short time. See the chapter, "Care
of the microscope" → p. 64, for additional in-
structions.
• Instructions and list of microscope default
settings (“Identification Sheet”)
The system box contains the microscope acces-
sories:
• Tube
• Eyepieces
Caution!
• Objectives
Do not yet connect the microscope and pe-
ripherals to the power supply at this point!
• Condenser
• Lamp housings with accessories
• Fitting tool
• Depending on configuration, additional micro-
scope accessories such as filter cubes, etc.
14
Download from Www.Somanuals.com. All Manuals Search And Download.
4. Unpacking the Microscope
Installation location
Transport
Work with the microscope should be performed For shipping or transporting the microscope
in a dust-free room, which is free of oil vapors and its accessory components, the original
and other chemical vapors, as well as extreme packaging should be used.
humidity. At the workplace, large temperature
fluctuations, direct sunlight and vibrations As a precaution to prevent damage from vibra-
should be avoided. These conditions can distort tions, the following components should be disas-
measurements and micrographic images.
sembled and packaged separately:
• Unscrew the objectives.
• Remove the condenser.
Allowable ambient conditions
Temperature
15-35°C
Relative humidity
maximum 80% up to 30°C
Microscopes in warm and warm-damp climatic • Remove the stage.
zones require special care in order to prevent
the build up of fungus.
• Remove the lamp housings.
See the chapter, "Care of the microscope" → p. 64,
for additional instructions.
• Disassemble the burner of 106 z lamp housing.
• Remove all moving or loose parts.
Caution:
Electrical components must be assembled at
least 10 cm from the wall and away from
flammable substances.
15
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
5. Assembling the Microscope
When using intermediate systems and optical
accessories, the sequence may vary.
In this case, read Chapter,
The microscope components are logically as-
sembled in this order:
"5.10 Optional accessories" → p. 29
• Stage
• Condenser
• Tube
• Eyepieces
• Objectives
• Light sources
• Filter cubes/reflectors*
Only a few commonly used screwdrivers and
keys are necessary for assembly, which are in-
cluded in the delivery package.
16
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
5.1 Stage
• From above, set the stage clamp onto the
dovetail guide (4.2) and push the stage down-
wards until the upper end of the dovetail guide
is tightly fastened to the upper end of the
stage clamp.
Caution:
!
Before assembling the stage, make sure no ob-
jectives are installed!
• Firmly tighten the stage clamp (4.1).
• Place the specimen holder on the stage and
fasten it with the two screws (3.1).
• Using the condenser height adjuster (3.2), turn
the condenser holder completely upwards, i.e.
as close to the stage as possible.
Note:
For thicker specimens (Leica DM4000 M) the
stage can be set to a correspondingly lower
level.
• Loosen the stage clamp (3.3) slightly.
Fig. 3
Mechanical object stage
Fig. 4
Assembling the stage
Stage clamp
Dovetail guide
1
2
3
Locking screws for specimen holder
Condenser height adjuster
Stage clamp
1
2
1
1
2
2
3
17
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
5.2 Condenser
Note:
The condenser must be centered before using
the microscope.
→ Köhler illumination p. 37.
• Using the condenser height adjuster (5.4), turn
the condenser holder (5.1) completely down-
wards.
• Unscrew the clamping screw for the con-
denser (5.3) far enough so that the condenser
can be inserted from the front.
Fig. 6
Underside of condenser
1
1
Orientation pin
• From the front, insert the condenser into the
condenser holder as far as it will go. On the
underside of the condenser, there is an orien-
tation pin (6.1), which must be located in the
guiding notch (7.1).
Fig. 7
Condenser holder
• Pull the condenser’s clamping screw (5.3) so
that the condenser is locked in place.
1
Guiding notch
• Connect the condenser over the connection
(8.1) with the stand.
1
Fig. 5
Condenser holder
1
2
3
4
Condenser holder
Condenser centering
Fig. 8
Condenser connector
Clamping screw for condenser
Condenser height adjuster
1
Condenser cable socket
1
2
3
4
1
18
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
Fig. 9
Fastening the tube
5.3 Tube and Eyepieces
1
Clamping screw
The tube is mounted to the stand either directly or
with the use of intermediate modules. It is fastened
in place with the side clamping screw (9.1).
1
• Loosen the clamping screw (9.1).
• Insert the tube in the circular receptacle
(dovetail ring).
• Retighten the clamping screw (9.1).
• Only for the MBDT motorized tube:
Connect the tube to the stand with the con-
nector socket (10.1).
Fig. 10 Motorized tube connection
Connector socket
• The eyepieces are inserted into the eyepiece
tubes on the tube.
1
6.4 Objectives
1
The receptacles on the objective turrets are
numbered (Fig. 11). The individual objectives
have already pre-assigned positions at the
factory according to their configuration.
A list of the exact objective positions is provided
in shipment. (“Identification Sheet”)
Caution:
!
Cover unoccupied threads on the turret with
dust protector caps!
Fig. 11
Objective turret
with labeled
objective
receptacles
19
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
Fig. 12
5.5 Light Sources for the Transmitted Light Axis
Lamp housing 107/2
Releasing the
fastening screw
Caution:
Be sure that the lamp housing is discon-
nected from the power supply. Unplug the
power plug and the power supply during as-
sembly.
107/2 Lamp Housing
This lamp housing is used with a 12V 100W halo-
gen lamp, which is already mounted.
In case the lamp has to be removed:
Fig. 13
1
Lamp housing 107/2,
opened
1
Mount with
halogen lamp
Collector
• Remove the fastener screw on the housing
(Fig. 12).
2
• Remove the housing by pulling it upwards.
• Remove the lamp
2
• Insert the new 12V 100W lamp (13.1) with dust
cover straight into the socket until it stops. Be
sure that the lamp is inserted straight.
• Remove the lamp’s dust cover.
Fig. 14 Rear side of stand
1
2
3
4
Incident light lamp housing receptacle
Caution:
Transmitted light lamp housing receptacle
12 V 100 W connection for transmitted light (symbol: ꢀ)
ꢀ
Do not remove the lamp’s dust cover until
you have installed the lamp. Avoid
fingerprints on the lamp.
12 V 100 W connection for incident light (symbol:
)
• Replace the housing and fasten it in place us-
ing the fastening screw.
1
2
• Place the lamp housing in the transmitted light
lamp housing receptacle (14.2) and fasten it
with the clamping screw on the side.
3 4
• Connect the lamp housing to the power
supplyfor transmitted light (symbol: ꢀ) (14.3).
20
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
5.6 Light Sources for the Incident Light Axis
Caution:
This lamp housing is used with a 12V 100W halo-
gen lamp or various gas discharge lamps.
Inserting the 12V 100W halogen lamp into the
106 z lamp housing
During assembly, always unplug the power
supply unit of the 106 z lamp housing from its
socket.
• Unscrew the fastening screws of the cover
and lift up the cover (16.1).
• Unscrew the fastening screws of the lamp
mount (16.8) and pull out the mount (Fig. 17).
Never touch the glass parts of the burner
with bare hands.
Never look directly into the beam path (blind-
ing hazard).
During assembly work on xenon burners, al-
ways wear the supplied protective gloves and
face protection (Fig. 15) (risk of explosion).
106 z lamp housing
Fig. 16 106 z lamp housing (on the side, open)
1
2
3
Cover raised
Collector
12 V 100 W lamp or
gas discharge lamp in mount
Reflector (mirror)
4
5, 6, 7 Adjusting screw for x-y reflector
8
9
Fastening screw for lamp mount
Socket for contact plug
Fig. 15
1
Protective gloves and mask
4
2
5
6
7
3
8
9
8
21
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
• Insert the lamp with the dust cover straight • Place the lamp housing in the incident light
into the socket until it stops.
lamp housing receptacle (18.1) and fasten it
with the clamping screw on the side.
• Remove the dust cover.
• Connect the lamp housing to the power supply
ꢀ
• Reinsert the lamp mount and retighten the fas-
tening screw (16.8).
for incident light (symbol ) (18.4).
Caution:
Do not remove the lamp’s dust cover until af-
ter you have installed the lamp. Be certain to
avoid getting fingerprints on the lamp.
• Close the lamp housing and retighten the fas-
tening screws.
Fig. 18 Rear side of stand
1
2
3
4
Incident light lamp housing receptacle
Transmitted light lamp housing receptacle
12 V 100 W connection for transmitted light (symbol: ꢀ)
ꢀ
12 V 100 W connection for incident light (symbol:
)
Fig. 17 Lamp mount with 12 V 100 W halogen lamp
1
2
3 4
22
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
Inserting the gas discharge lamps (Hg and Xe)
into the 106z lamp housing
Hg and Xe lamps are powered by the separate
ebq 100 supply unit.
Read the separate instruction manual provided
with this supply unit.
The following gas discharge lamps may be used
and require different lamp mounts (Fig. 19):
Type
Typical bulb life*
50 W high-pressure mercury burner (alternating current)
100 W high-pressure mercury burner (direct current, stabilized/not stabilized
100 hrs.
200 hrs.
100 W high-pressure mercury burner (direct current, stabilized/not stabilized, type 103 W/2) 300 hrs.
75 W High-pressure xenon burner (direct current, stabilized)
400 hrs.
* Please regard the data sheets of the burners.
23
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
• To open the 106 z lamp housing, unscrew the
fastening screws on the cover.
Caution:
• Remove the transport anchorage (red plastic
rod in place of the burner) in the lamp mount.
To do so, remove the lower clamp (19.1). Pull
up the cooling element (19.3) and turn it to the
side. Detach the lower clamp system (19.2)
and remove the transport anchorage.
Hg 50 burner:
After installation, the labeling must be upright.
If a glass melt nipple is present (19a.4), posi-
tion it by turning the burner so that the nipple
does not come in the way of the beam path
later, but instead is positioned sideways.
• Install the burner in mirror image fashion.
Xe 75 burner:
Remove the burner’s dust cover (19b.5) after
you have installed the burner.
Fig. 19 a-d Lamp mounts for gas discharge lamps
1
4
Upper clamping system, 2 Lower clamping system, 3 Cooling element
Nipple of the mercury 50 burner, 5 Dust cover of the mercury 75 burner
Hg 50
Xe 75
a
b
3
5
3
2
1
2
1
4
Hg 100
c
Hg 100
d
3
Stab.
3
1
1
2
2
24
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
• Insert the lamp mount, with the burner in-
stalled, into the lamp housing and tighten it
with the screws (20.8).
• Put the lid down again. Plug in the contact
plug as far as it goes and retighten the
screws.
Fig. 20 106 z lamp housing (on the side, open)
1
2
3
Cover raised
Collector
12 V 100 W lamp or
gas discharge lamp in mount
Reflector (mirror)
• Place the lamp housing in the incident light
lamp housing receptacle (21.1) and fasten it
with the clamping screw on the side.
4
5, 6, 7 Adjusting screw for x-y reflector
8
9
Fastening screw for lamp mount
Socket for contact plug
• Connect the lamp housing to the power supply
(22.1).
1
4
2
5
3
6
7
Fig. 21 Rear side of stand
1
2
3
4
Incident light lamp housing receptacle
Transmitted light lamp housing receptacle
8
9
8
12 V 100 W connection for transmitted light (symbol: ꢀ)
ꢀ
12 V 100 W connection for incident light (symbol:
)
Fig. 22 Rear side of the ebq 100 supply unit
1
Lamp connection
1
1
2
3 4
25
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
Fig. 23 Filter cube
Fig. 24 Filter cube
5.7 Equipping the Incident Light filter turret
front side
back side
The receptacles on the turret are numbered.
According to your equipment, the individual filter
and/or reflector cubes have already pre-
assigned positions. A list is provided along with
your shipment (“Identification Sheet”).
Insert the filter and reflector cubes in the follow-
ing manner:
• Equip the incident light turret only when the
microscope is switched off.
Fig. 25 Removing the front panel
1
2
3
Filter receptacle
Retention pin
Front panel
• Remove the face plate from the upper part of
the microscope (Fig. 25). Turn the turret in any
direction until the locking pin engages.
• Insert the filter or reflector cube into the
mounting in front of you according to the
identification sheet provided.
To do so, place the filter or reflector cube on
the right side and press it to the left into the
mounting (Fig. 26).
1
2
3
• Push the retention pin (25.2) and continue to
turn the filter turret until you reach the next
locking position.
• Again make sure that the turret engages
(retention pin unlocks) and insert the next
filter and/or reflector cube as described
above.
Fig. 26 Inserting the filter or reflector cubes
Mounting
1
1
• When all filters and reflector cubes have been
inserted, close the front cover plate again.
5.8 Polarizer and Analyzer
26
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
Fig. 27 Assembly of the ICT/P transmitted light polarizer
ICT/P transmitted light polarizer
1
Clamping screw
• Using the left clamping screw, fasten the ICT/P
transmitted light polarizer to the underside of
the condenser holder (Fig. 27).
• Make sure that the red index point on the front
of the polarizer is aligned with 0.
• If necessary, insert the compensators (λ- and
λ/4 plates) into the polarizer’s receptacle
(Fig. 28).
1
Fig. 28 Inserting the compensators
Incident light polarizers:
R/P polarizer, rotating polarizer
L/ICR, R/ICR polarizer
• Remove the plug cap on the right side of the
incident light axis (Fig. 29).
Fig. 29 Inserting the polarizer
1
The plug cap is replaced with the polarizer.
• Insert the polarizer into the receptacle until it
latches in place.
Motorized polarizer
1
• A motorized polarizer is already installed and
ready for operation in the DIC condenser.
Transmitted light and incident light analyzer
27
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
• Remove the plug cap on the left side of the • Insert the objective prism into the tube slot
stand.
(Fig. 31.1). The code letter must match the
code letter on the objective.
• Insert the polarizer into the receptacle until it
latches in place (Fig. 30).
• With the microscope Leica DM5000 B the DIC
objective prisms are already mounted in the
DIC turret above the objective revolving
nosepiece(Fig. 68).
Motorized analyzer
• Insert the analyzer cube as described in sec-
tion 5.7 "Equipping the Incident Light filter
turret" → p. 26, in the corresponding position
on the filter turret. See the list provided
(“identification Sheet”) for the correct
position.
5.10 Optional Accessories
6.9 DIC Prisms
Fig. 31 Inserting the objective prism slide
Fig. 30 Inserting the analyzer
1
Objective prism slide
1
The plug cap is replaced with the analyzer.
1
1
28
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
Ergomodule
• Remove the clamping ring from the filter slide.
For raising the eye level of the tube opening, the
ergomodule may be used.
It is fastened in place with the side clamping
screw.
• Insert the Booster Lens or Excitation
Manager.
• Push the cover back.
• Insert the clamping ring.
Mirror Housing
• Place the mirror housing directly onto the
lamp housing receptacle on the back of the
stand and attach it using the side clamping
screw.
• Insert the filter slide into the front receptacle
on the right side of the stand (32.1, 33.1).
• Using two filter sliders, the Excitation
Manager can be inserted into the back
receptacle.
• Place the lamp housing onto the mirror hous-
ing and fasten it using the corresponding
clamping screw on the side.
5.11 Connection to the Power Supply
Booster Lens / Excitation Manager
Fig. 32
Fig. 33
1
Insert of Booster Lens
1
Insert of Excitation Manager
1
1
29
Download from Www.Somanuals.com. All Manuals Search And Download.
5. Assembly
After completing the assembly work, connect Only for the Leica DM5000 B:
the stand to the power supply using the power
cable (Fig. 34.2).
• Connect the microscope (36.1) to the
5.12 Connection to the CTR5000 Electronics Box
"Microscope" jack (35.1) on the rear of the
electronics box. Use the cable with the 25-pin
plug.
Fig. 34 Rear side of stand Leica DM4000 B/M
1
2
Power switch
Power supply
• Connect the electronics box to the power sup-
ply using the power cable (35.2).
1
2
Fig. 35 Rear side of electronics box CTR5000
Fig. 36 Rear side of stand Leica DM5000 B
1
2
Microscope connection
Power supply
1
Connection to the CTR5000 electronics box
1
2
1
2
30
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
6. Startup
6.1 Functional Principle
The microscope’s most important functions may be easily accessed using function keys.
• The microscope may be switched between various contrast processes by pressing a single button.
• The microscope recognizes the objective chosen and the respective contrast process. There-
fore, the values for intensity (INT), aperture diaphragm (AP) and field diaphragm (FD) are al-
ways set correctly.
• The values for INT, AP and FD can be changed individually. This overwrites the previous
setting. Actual settings are stored automatically.
• The specifications for INT, AP and FD always relate to the currently activated light axis (trans-
mitted light or incident light).
• In addition to the preset function keys for INT, AP and FD, there are also variable function keys.
Variable function keys:
• These function keys are assigned logical functions before delivery(see “Identification Sheet”)
• These functions can be reprogrammed according to your individual wishes.
Note: (Reset-Function)
The microscope can be reset to the default functions programmed at the factory:
• When the microscope is switched off, press all 3 variable function keys on the left stand
section.
• Switch on the stand.
• Hold the keys pressed down until initialization is complete.
• The standard information is shown in the display.
• Switch off the instrument and switch it on again. The settings are stored now.
6.2 Switching on the Microscope
• First, swivel the objective with the least magnification into position.
31
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
Possible Assignments for the Function Keys
For Leica DM4000 B/DM5000 B:
Function key
Meaning
BF
Bright field (Transmitted light)
PH
ICT
DF
Phase contrast (Transmitted light)
Interference contrast (Transmitted light)
Dark field (Transmitted light)
POL
Polarization (Transmitted light)
|
CHANGE TL
Switch through all transmitted light processes
INT ↑
INT ↓
Increase brightness (transmitted light)
Reduce brightness (transmitted light)
Open aperture diaphragm (transmitted light)
Close aperture diaphragm (transmitted light)
Open field diaphragm (transmitted light)
Close field diaphragm (transmitted light)
AP
↑
AP ↓
FD
FD
↑
↓
SHUTTER TL
Open/close transmitted light shutter
FLUO
Fluorescence (last filter cube)
CUBE 1
Select fluorescence cube at position 1
|
|
CHANGE CUBE
CHANGE CUBE
SHUTTER FLUO
Switch through fluorescence cubes in clockwise fashion
Switch through fluorescence cubes in counterclockwise fashion
Open/close fluorescence shutter
INT FLUO ↑
INT FLUO ↓
FD FLUO ↑
FD FLUO ↓
Increase brightness (fluorescence)
Reduce brightness (fluorescence)
Open field diaphragm (fluorescence)
Open field diaphragm (fluorescence)
|
COMBI
Combination mode
(PH / fluorescence or ICT / fluorescence)
CHANGE COMBI | Switch through all combination modes
32
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
For Leica DM4000 M:
Function key
Meaning
BF
Bright field (Incident light)
ICR
DF
Interference contrast (Incident light)
Dark field (Incident light)
POL
Polarization (Incident light)
CHANGE RL
Switch through all incident light processes
|
INT ↑
INT ↓
Increase brightness (incident light)
Reduce brightness (incident light)
AP
AP ↓
FD
FD
↑
Open aperture diaphragm (incident light)
Close aperture diaphragm (incident light)
Open field diaphragm (incident light)
Close field diaphragm (incident light)
↑
↓
SHUTTER RL
Open/close incident light shutter
FLUO
Fluorescence (last filter cube)
CUBE 1
Select fluorescence cube at position 1
Switch through fluorescence cubes
CHANGE FLUO
FOCUS FINDER
Select smallest field diaphragm and switch back to original field
diaphragm by pressing the key again
BF TL
INT ↑
INT ↓
Bright field (Transmitted light )
Increase brightness (transmitted light)
Reduce brightness (transmitted light)
Open aperture diaphragm (transmitted light)
Close aperture diaphragm (transmitted light)
Open field diaphragm (transmitted light)
Close field diaphragm (transmitted light)
AP
↑
AP ↓
FD
FD
↑
↓
COMBI |
Combination process (BF and BF TL)
33
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
• Switch-on the microscope at the power
switch (34.1,36.1). All motorized microscope
components first undergo an initialization
phase.
After turning on the gas discharge lamps, the
burner must be immediately adjusted. There-
fore, do not turn on the power supply unit
yet. First, work in transmitted light in order to
familiarize yourself with the microscope’s
controls.
After initialization is complete, the display on the
stand shows the current microscope setting (Fig.
37).
6.3 The Display (Leica DM4000 B/DM4000 M)
The microscopic components such as dia-
phragms, condenser, light and phase rings are
already pre-centered in the factory. However,
re-centering may be necessary due to transpor-
tation and assembly.
Before proceeding with the necessary steps,
first familiarize yourself with the stand’s display
and control panel.
Caution:
Fig. 37 Display after initialization
34
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
The display shows the current microscope set- ꢂ ꢃ
tings. The display depends on the microscope’s
Light Intensity
configuration. In the first column, corresponding
pictograms indicate the type of information: con-
trast method, magnification, light intensity, dia-
phragms, light splitting for photo tubes.
Please see the abbreviation index for a list of ab-
breviations and pictograms used → p. 67.
The actual brightness setting is graphically de-
picted by a beam. Additionally, the light intensity
is indicated in 20 (coarse adjustment) or in 255
(fine adjustment) increments → p. 52.
ꢂ ꢃ
Diaphragms
Contrast Method
The values for the field diaphragm (FD) and the
aperture diaphragm (AP) are indicated numeri-
cally. The field diaphragm may be either round or
rectangular. Accordingly, the FD designation is
set in parentheses or in brackets: (FD) or [FD].
In the first row, you find an indication of the ac-
tive light axis (transmitted light or incident light)
of the current contrast method and the current
filter cube.
Note:
The shutter status is displayed for the
transmitted light or incident light shutter:
When using a digital camera, rectangular field
diaphragms are recommended.
Transmitted light shutter open
↑
↑
Transmitted light shutter closed
Beam splitting
Incident light shutter open
Incident light shutter closed
↓
↓
If a motorized tube is used, the light splitting
between ocular (Eye) and photo output (Docu) is
indicated in %.
Magnification
+
The current objective magnification, sometimes
followed by the re-magnification of the magnifi-
cation changer, appears along with the total
magnification:
Note:
The display may flash after the initialization
phase or even during microscopy session. This
always occurs when the contrast method
selected can not be performed with the actual
microscopic settings. For example, an objective
may be swiveled in that is not suited to the
contrast method chosen.
Σ = Objective x Re-magnification x Eyepiece
6.4 The Function Keys
Then check your settings.
35
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
There is a row of function keys both on the right
and left side of the stand. Some of these keys Variable function keys:
are defined, and some of them are variable. The A factory preset is performed which fits your
variable function keys have various meanings microscope configuration. The function keys are
depending on the microscope configuration.
labeled accordingly, and a separate description
of the key occupation accompanies the
Defined Function Keys on the left side of the microscope (“Identification Sheet”).
stand
The TL/IL key (38.1) switches between incident Abbreations are listed on p.32f.
light and transmitted light. The last contrast
method used is restored.
The INT (38.3) keys adjust the light intensity indi- 6.5 Köhler Illumination
vidually. Settings can be made either in large or
For each objective, optimal values for the aper-
small increments. Pushing both INT buttons at
the same time switches between coarse and
fine setting. The display indicator changes
accordingly → p. 52.
ture diaphragm and the field diaphragm are al-
ready set. The condenser is also already
adjusted in the factory.
The AP (38.4) keys for the aperture diaphragm
and FD (38.2) for the field diaphragm are used
to set each diaphragm. The optimal values are
automatically preset when selecting the con-
trast method.
Fig. 38 Defined Function Keys
1
2
3
4
Transmitted light/incident light
Field diaphragm
Light Intensity
Aperture diaphragm
3
2
4
1
36
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
However, depending on how the condenser is • Insert the specimen in the stage’s specimen
disassembled and reassembled, it may be nec-
essary to re-adjust the condenser in some
holder (39.3).
cases. Therefore, check the condenser • Focus on the specimen. The focus wheel on
centering.
the left side of the stage allows focus adjust-
ment in large and small increments. On the
right side of the stage, there is also a focus
wheel for fine focus adjustment.
The following procedure is provided for the
transmitted light-bright field illumination.
• Select an objective with moderate • Set the light intensity using the INT keys (38.3).
magnification (10x-40x).
• Close the field diaphragm with the FD function
• Activate the transmitted light axis by pushing
the TL/IL button (38.1). "TL" appears in the first
line of the display.
key (38.2) until the edge of the diaphragm ap-
pears in the specimen plane.
• Using the condenser height adjuster (39.4), ad-
just the condenser until the edge of the field
diaphragm appears in sharp relief.
• If the image does not appear in the middle of
the field of view (41c), the condenser must be
moved into the middle of the field of view with
the help of the two leveling screws (40.1).
• Choose "bright field" as the contrast method by
pressing the BF (one of the variable function
keys, behind the focus dials).
"TL BF" appears in the first line of the display.
Fig. 39 Stage with specimen holder
1
2
3
4
Object motion (X direction)
Object motion (Y direction)
Specimen holder
Condenser height adjuster
3
2
1
4
37
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
• Open the field diaphragm just enough for it to Do not adjust the aperture diaphragm. The aper-
disappear from the field of view (41d).
ture diaphragm is already set optimally for each
objective.
6.6. Checking Phase Contrast Rings
Caution:
Fig. 40 Condenser centering
Fig. 41 Köhler Illumination
1
Centering bolts
a
b
c
Field diaphragm not focused, not centered
Field diaphragm focused, but not centered
Field diaphragm focused and centered
Diameter is too small, however
d
Field diameter (light) = Field diameter (view)
(Köhler Illumination)
A
C
B
D
1
1
38
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
If your microscope is equipped for the use of
phase contrast, the light rings that fit the objec- • In the place of an eyepiece, insert the focus-
tives are built into the condenser.
The light rings are already leveled in the factory.
However, the leveling should be rechecked.
ing telescope (Fig. 42) into the observation
tube.
• Swivel in the phase contrast objective with
the least magnification.
Note:
• Focus the ring structure (43a) by slightly loos-
ening the clamping ring (42.2) and moving the
eyelens (42.1).
Every objective is assigned its own light ring in
the condenser disc. Therefore, a check must be
performed for each objective. When swiveling in
a suitable objective for phase contrast, the cor-
responding light ring is set automatically.
• Retighten the clamping ring.
• Press the PH (Phase Contrast) button. The ring
diaphragm in the condenser is pivoted in.
• Press the BF (Bright Field) button (one of the
variable function keys, behind the focus dials).
• If the light ring and the phase ring are not
shown as arranged in Fig. 43c, the light ring
must be leveled.
Fig. 42 Focusing telescope
Fig. 43 Phase contrast centering procedure
1
2
Adjustable eyelens
PH=phase contrast ring, LR=light ring
Clamping ring for fixing the focus position
a
b
Condenser in bright field (BF) position
Condenser in phase contrast (PH) position
Light ring (LR) not centered
c
Light ring and phase ring centered
A
B
C
1
2
PH
LR
39
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
Fig. 44 Light ring centering
• Insert the centering key through the corre-
sponding openings (44.1) in the condenser
holder.
1
Clamping screw
• Turn the centering screws until the dark ring
(phase ring in the objective) is congruent with
the slightly narrower bright ring (light ring in
condenser) (43 c).
1
• Repeat the process for all other phase con-
trast objectives.
• Remove the centering keys after the centering
procedure.
Note:
During change of objectives the centering keys
must not remain in the openings of the
condenser.
6.7 Adjusting the Light Sources
Transmitted Light Axis (TL) with 107/2 Lamp
housing
40
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
The 107/2 lamp housing with 12 V 100 W halogen
lamp has a defined presetting. The lamp need
not to be centered.
When switching to the BF or Smith reflectors,
there is a danger of being glared!
For the 106 z lamp housing, the direct filament im-
age (for halogen lamps) or direct arc image (for gas
discharge lamps), and its mirror image are focused
Incident light axis (IL) with 106 z lamp housing
• When a supply unit is used, it is turned on first. separately and adjusted to each other.
• Activate the incident light axis using the TL/IL On the left side of the microscope, there is an
function key. FLUO (Leica DM4000 B/ DM5000 B) adjustment window (1.14, p. 12) for mapping the
or IL (Leica DM4000 M) appears in the display. light source.
• Insert the reflector for lamp adjustment While observing the light source in the adjust-
(Fig. 45) into the filter turret in place of a filter ment window, the lamp is adjusted as follows:
cube. (See → p. 26).
Note the name of the exchanged filter cube.
Centering the 12 V 100 W Halogen Lamp
• Turn the reflector into the light path.
The reflector has reached the correct position
when the name of the exchanged filter cube is
shown in the upper right of the display.
Fig. 46 106 z lamp housing
1
Lamp height adjustment
2,4 Mirror image height and side adjustment
Caution:
3
5
6
Focusing the reflector
Lamp side adjustment
Never look directly into the light path!
Collector (focusing of the lamp image)
5
1
6
Fig. 45 Reflector cube for lamp adjustment
(similar to illustration)
2
3
4
41
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
Fig. 47 Direct lamp filament image focused,
but not centered
• In the adjustment window, you see the direct
filament image and the mirror image, which in
most cases are shifted together.
(in reality, the image is less focused)
• Focus the direct filament image with the col-
lector (46.6).
• Use the adjusting buttons on the rear side of
the lamp housing (46.2, 46.4) to pivot the lamp
filament’s mirror image to the side or com-
pletely out of the beam path. The lamp fila-
ment’s focused image remains visible (Fig. 47).
• Adjust the direct filament image using the ad-
justing knobs (46.1) and (46.5) so that the
centering surface is halfway covered (Fig. 48).
Fig. 48 Direct lamp filament image in target position
(in reality, the image is less focused)
• Then pivot the lamp filament’s mirror image
with the adjusting knobs (46.2 and 4), and
focus it using the reflector (46.3).
• Align the mirror image symmetrically to the fil-
ament image (Fig. 49). To do so, use the adjust-
ing knobs (46.2) and (46.4) again.
• Defocus the image with the collector head
(46.6) until the filament image and mirror im-
age are no longer recognizable and the image
is uniformly illuminated.
• Exchange the reflector cube for lamp
adjustment for the original filter cube.
Note:
Turn off the microscope before exchanging
the reflector cube.
Fig. 49 Direct lamp filament image and mirror image in
target position
(in reality, the image is less focused)
Centering the Hg 50 W mercury lamp
• In the adjustment window, you see the direct
arc image and the mirror image, which in most
cases are shifted together.
42
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
Fig. 50 Direct arc image focused but decentered
(in reality, the image is less focused)
• Focus the direct image with the collector
(46.6).
• Use the adjusting buttons on the rear side of
the lamp housing (46.2,46.4) to pivot the arc’s
mirror image to the side or completely out of
the beam path. The lamp filament’s focused
image remains visible (Fig. 50).
• Use the adjusting buttons (46.1) and (46.5) to
place the direct arc image right or left on an
imaginary center line of the centering plane
(Fig. 51).
• Then pivot the arc’s mirror image with the ad-
justing knobs (46.2 and 4) and focus it using
the reflector (46.3).
Fig. 51 Direct arc image in target position
(in reality, the image is less focused)
• Use the adjusting knobs (46.2 and 4) to orient
the mirror image symmetrically to the direct
image (Fig. 52).
• Defocus the image with the collector knob
(46.6) until the arc image and mirror image are
no longer recognizable and the image is
uniformly illuminated.
• Exchange the reflector cube for lamp
adjustment for the original filter cube.
Fig. 52 Direct arc image and mirror image in target
Centering the Hg 100 W and Xe 75 W
mercury lamps
position (in reality, the image is less focused)
• In the adjustment window, you see the direct
arc image and the mirror image, which in most
cases are shifted together.
43
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
Fig. 53 Direct arc image focused but not centered
(in reality, the image is less focused)
• Focus the direct image with the collector
(46.6).
• Use the adjusting buttons to pivot the arc’s
mirror image on the rear side of the lamp
housing (46.2,46.4) to the side or completely
out of the beam path. The arc’s focused im-
age remains visible (Fig. 53).
• Use the adjusting buttons (46.1 and 5) to place
the direct arc image in the middle of the
centering plane, whereby the bright tip of the
arc, the focal spot, should lie slightly outside
the center (Fig. 54).
Fig. 54 Direct arc image in target position
• Then pivot the arc’s mirror image with the ad-
justing knobs (46.2) and (46.4) and focus it us-
ing the reflector (46.3).
(in reality, the image is less focused)
• Use the adjusting knobs (46.2 and 4) to orient
the mirror image symmetrically to the direct
image (Fig. 55).
The V-shaped irradiation of the direct image
and mirror image arcs can be superimposed.
Caution:
The bright tips of the arcs, the focal spots, must
never be projected onto each other, as this re-
sults in a danger of explosion by overheating.
Fig. 55 Direct arc image and mirror image in target
position (in reality, the image is less focused)
44
Download from Www.Somanuals.com. All Manuals Search And Download.
6. Startup
In older lamps, the structure of the arc is no
longer clearly recognizable. The image is
then more like that of a HG 50 lamp. The im-
age and mirror image can no longer be su-
perimposed exactly. In this case, align both
images.
• Using the collector, defocus the image with
the knob (46.6) until the arc image and mirror
image are no longer recognizable and the im-
age is uniformly illuminated.
• Exchange the reflector cube for lamp
adjustment for the original filter cube.
Note:
Turn off the microscope before exchanging
the reflector cube.
45
Download from Www.Somanuals.com. All Manuals Search And Download.
7. Operation
7. Operation
7.1 Switching on the Microscope
7.2 Stages and Specimen Displacement
Lengthening the coaxial pinion
When using a gas discharge lamp, the ebq 100
external supply unit must be turned on
separately (56.1).
Then switch-on the microscope at the power
switch.
All motorized microscope components first un-
dergo an initialization phase.
• For lengthening, pull the lower grip (58.2)
downwards. Repeat with the upper grip (58.1).
After the initialization is complete, the display on
the stand (Fig. 57) shows the current microscope
setting.
Torque adjustment
The torque is already optimally set at the factory,
however, it can be individually adjusted using
two knurled rings (58.3, 58.4).
Fig. 56 Front view of the ebq 100 supply unit
1
2
Power switch
Lamp status
Fig. 58 Revolving object stage
1
2
3
4
5
Object motion (Y direction)
Object motion (X direction)
Torque adjustment (Y direction)
Torque adjustment (X direction)
Focus dial for fine focusing
1
2
Fig. 57 Display after initialization
3
5
1
2
4
46
Download from Www.Somanuals.com. All Manuals Search And Download.
7. Operation
Rotating the Stage
The swiveling range of the rotating stages is There is a focus dial on the left side of the stage for
0°- 110°. coarse and fine focus adjustment (Fig. 59).
7.3 Focusing
• In order to revolve the stage, loosen the fas- On the right side of the stand, there is also a
tening screw (59.1).
focus dial, which is used exclusively for fine
focusing (58.4).
• Bring the table into the desired position.
• Retighten the fastening screw.
The special design of this dial makes it possible
to simultaneously grasp the coaxial drive with
your hand while operating the fine drive with
one finger.
Fig. 59 Revolving object stage
1
2
3
Clamping screw
Fine focusing
Coarse focusing
1
2
3
47
Download from Www.Somanuals.com. All Manuals Search And Download.
7. Operation
7.4 Tubes
Adjusting the Viewing Angle
• For the AET22 and EDT22 ergotubes, the view-
ing angle can be adjusted by tilting the binocu-
lar viewer in the range of 5° - 32° (Fig. 61).
Note:
Close any unused tube openings, as otherwise
stray light can interfere with observation.
Adjusting the Eyepiece Extension to the Arm
Length
Note:
• With the AET22 tube, the eyepieces can be
extended up to 30 mm (Fig. 61).
Make sure that the connector cable is plugged in
on the MBDT25+ motorized tube (60.1).
Adjusting the Viewing Distance
• Adjust the viewing distance of the eye-
pieces so that a congruent total image is seen
(Fig. 60).
Fig. 60 Tube setting
Fig. 61 With AET22 tube individual adjustments
↔
1
Personal eyebase settings
Motorized tube connection
↔
1
48
Download from Www.Somanuals.com. All Manuals Search And Download.
7. Operation
Beam Splitting in Photo Tubes
7.5 Eyepieces
EDT22 tube:
The beam splitting between the observation and
documentation outputs has a definite presetting
(50:50).
Note:
The eyepiece’s aperture protector must be
removed,or at least folded back, during
microscopy while wearing eyeglasses.
Eyeglasses with multifocal lenses (bifocals and
smooth view glasses) must be removed while
operating the microscope.
BDT25+ tube:
The beam splitting is set manually by pulling out
a control bar.
Control Bar
VIS
Observation
100 %
Photo
0 %
50/50
PHOTO
150 %
110 %
50 %
100 %
• For the adjustable tubes with documentation
output, choose the 100% position.
MBDT25+ tube:
This tube is similar to the documentation tube
BDT25+, but it is motorized.
Eyepieces with Inlaid Reticle
•
•
•
Focus the reticle by adjusting the eyelens.
Focus on the object through this eyepiece.
The control positions are selected using a vari-
able function key on the stand.
HC L 2TU tube:
The beam splitting is set manually by pulling out
a control bar.
Then, close that eye and focus on the
specimen by adjusting only the second
ocular.
Control Bar
VIS
Observation
100 %
Photo
0 %
Correction for Vision Problems
PHOTO
110 %
100 %
•
With your right eye, look through the right
eyepiece and bring the specimen into sharp
focus.
Fig. 62 BDT25+ tube with digital camera
Control bar
1
•
Then, with your left eye, view the same speci-
men and rotate the left eyepiece tube until
the object is brought into sharp focus. Do not
use the focus dial.
1
49
Download from Www.Somanuals.com. All Manuals Search And Download.
7. Operation
• Start with a small level of magnification. Then
switch to the next higher objective.
7.6 Objectives
The objective must be moved manually into the
light path. Be sure that the nosepiece turret
locks into place.
• For immersion objectives use the appropriate
immersion medium.
OIL: only use optical immersion oil
according to DIN/ISO standards.
Cleaning → p. 65.
The objective’s position in the turret is factory-
set and must be adhered to while screwing in
the objectives (see Objective Assembly → p. 19)
W:
Water immersion.
IMM: Universal objective for water, glycerol,
oil immersion.
When you rotate the objective into position, the
microscope automatically recognizes:
Caution!
• the selected contrast method
Follow safety instructions for immersion oil!
• the optimal settings for field and aperture
diaphragm
• the optimal condenser setting
The objective magnification and the total magni-
fication appear in the display → p. 35.
50
Download from Www.Somanuals.com. All Manuals Search And Download.
7. Operation
For lockable immersion objectives:
7.7 Magnification Changer
• Lock these by pushing the front part upwards Optionally, a coded magnification changer can
until it stops (approx. 2 mm).
be used, which is manually operated.
On the knurled ring, the following magnification
• Then, after a gentle turning motion to the right, factors can be set:
the objective is locked (Fig. 64).
B Stand
1x
M Stand
1x
For objectives with corrective mounts:
1.25x
1.6x
1.5x
2x
• Turn the knurl to adjust the objective to the
thickness of the cover glass.
The selected factor is indicated in the display
and included in the total magnification.
Fig. 63 Immersion objective (released)
Fig. 64 Immersion objective (locked)
51
Download from Www.Somanuals.com. All Manuals Search And Download.
7. Operation
7.8 Light Sources
7.9 Aperture Diaphragm and Field Diaphragm
• The brightness is set using the function keys Both diaphragms are already factory-set to the
(65.5). Then, the INT function keys are as- optimum setting for the current objective.
signed to the currently active axis for trans-
mitted light (TL) or incident light (IL).
• The AP (65.2) keys for the aperture diaphragm
and the FD keys (65.4) for the field diaphragm
may be used to change each diaphragm’s set-
ting at any time.
Then, the function keys are assigned to the
currently active axis for transmitted light (TL)
or incident light (IL).
• For TL and IL:
Settings can be made either in large or small
increments. Pushing both INT buttons
simultaneously switches between coarse and
fine setting. The display indicator changes ac-
cordingly.
0-20
Coarse adjustment:
======
Caution:
When doing so, old values are overwritten and
the new values are stored!
0-255
Fine adjustment:
• For Fluo:
----------
Caution:
The brightness is set in 5 fixed steps (FIM):
100% / 55% / 35% / 20% / 10%
While using PH or DF the aperture diaphragm is
completely opened and locked.
Fig. 645 Control panel
1
2
3
4
5
Variable function keys
Aperture diaphragm
Transmitted light/incident light
Field diaphragm
Light intensity
1
2
3 4
5
52
Download from Www.Somanuals.com. All Manuals Search And Download.
8. Imaging Procedure for Leica DM4000 B/DM5000 B
8. Imaging Procedure
for Leica DM4000 B/ Leica DM5000 B
8.1 Transmitted Light
8.1.2 Phase Contrast
8.1.1 Bright Field (TL)
• Switch to the transmitted light axis (TL) by
pushing the TL/IL button.
• Switch to the transmitted light axis (TL) by
pushing the TL/IL button.
• Select the PH contrast (phase contrast) method.
Do so by pressing the PH variable key.
Alternatively: Press the CHANGE TL |
variable key.
(For key occupation please see “Identification
Sheet”.)
• Select the BF (bright field) contrast method.
Do so by pressing the BF variable key.
Alternatively: Press the CHANGE TL |
variable key.
(For key occupation please see “Identification
Sheet”.)
The display indicates PH.
The display indicates BF.
• Insert a transmitted light specimen.
• Insert a transmitted light specimen.
• Rotate an appropriate objective into place.
Objectives that are suitable for phase contrast
are engraved with PH.
• Rotate an appropriate objective into place.
• Bring the image into focus using the focus dial
and set the brightness using the INT function
key.
• Bring the image into focus using the focus dial
and set the brightness using the INT function
key.
53
Download from Www.Somanuals.com. All Manuals Search And Download.
8. Imaging Procedure for Leica DM4000 B/DM5000 B
8.1.3 Dark Field (TL)
Notes:
• Switch to the transmitted light axis (TL) by
pushing the TL/IL button.
• The microscope automatically selects the
correct light ring in the condenser.
• Select the DF (dark field) contrast method.
Do so by pressing the DF variable key.
• When selecting the phase contrast method,
the aperture diaphragm is opened completely
and may not be adjusted. To avoid errors in
operation, the function keys for setting the ap-
erture diaphragm (AP) are locked.
|
Alternatively: Press the CHANGE TL
variable key.
(For key occupation please see “Identification
Sheet”.)
The display indicates DF.
The dark field ring (dark field stop) is set au-
tomatically.
• Insert a transmitted light specimen.
• Rotate an appropriate objective into place.
• Bring the image into focus using the focus dial
and set the brightness using the INT function
key.
Notes:
• The maximum objective aperture which may
be used for dark field is 0.75. All objectives
with greater aperture are automatically
blocked for this procedure ("DF" flashes in the
display).
• The microscope automatically selects the
correct light ring in the condenser.
• When selecting the dark field method, the
aperture diaphragm is opened completely and
may not be adjusted. To avoid errors in opera-
tion, the function keys for setting the aperture
diaphragm (AP) are locked.
54
Download from Www.Somanuals.com. All Manuals Search And Download.
8. Imaging Procedure for Leica DM4000 B/DM5000 B
8.1.4 Polarization (TL)
• Bring the polarizer and analyzer into cross po-
sition until they reach maximum darkness.
• Switch to the transmitted light axis (TL) by
pushing the TL/IL button.
• Insert a specimen and rotate a suitable objec-
tive into place.
• Select the POL (polarization) contrast method.
Do so by pressing the POL variable key.
|
Alternatively: Press the CHANGE TL
Motorized procedure:
variable key.
(For key occupation please see “Identification
Sheet”.)
• After selecting the POL contrast method, the
condenser automatically switches to the posi-
tion of the polarizer. The analyzer cube is also
automatically brought into the light path.
The display indicates POL.
Combined procedure:
Mechanical procedure:
• For the Leica DM4000 B and Leica DM5000 B
microscopes, it is possible to combine
mechanical and motorized components.
• Turn the polarizer on the underside of the
condenser in the light path (Fig. 66). Make sure
that the red index point on the front of the
polarizer is aligned with 0.
• Insert the analyzer into the left side of the
stand (67.1).
Fig. 66 Swivel in polarizer
Fig. 67 Insert analyzer
1
Polarizer
1
Analyzer
1
1
55
Download from Www.Somanuals.com. All Manuals Search And Download.
8. Imaging Procedure for Leica DM4000 B/DM5000 B
8.1.5 Differential Interference Contrast (TL)
Alternatively:
• Manually rotate the polarizer on the underside
of the condenser into the light path (Fig. 66).
(only for DM5000 B)
• Switch to the transmitted light axis (TL) by
pushing the TL/IL button.
• Likewise, manually insert the analyzer into the
left side of the stand (Fig. 67).
Objective and coindenser prisms are
automatically moved into the light path as
well.
• Insert a specimen and rotate a suitable objec-
tive into place.
• Select the DIC contrast method.
Do so by pressing the DIC variable key.
Alternatively: Press the CHANGE TL
• Fine adjustment is possible using the knurled
ring above the objective nosepiece.
|
variable key.
(For key occupation please see “Identification
Sheet”.)
The display indicates ICT.
• The polarizer located in the condenser and the
fitting condenser prism are automatically
brought into the light path. The corresponding
objective prism and the analyzer cube are also
positioned automatically.
• For fine adjustment use the knurled ring above
the objective nose piece (Fig. 68).
Fig. 68 Objective prism slide
1
Knurled wheel for fine adjusting
1
56
Download from Www.Somanuals.com. All Manuals Search And Download.
8. Imaging Procedure for Leica DM4000 B/DM5000 B
8.2 Fluorescence
• The fluorescence intensity can be increased
using the Booster Lens on the right side of the
stand (Fig. 69).
• Switch to the fluorescent light axis (FLUO) by
pushing the TL/IL button.
• For multifluorescence, use of a Excitation
Manager is recommended. The Excitation
Manager is inserted into the right side of the
stand up to the last stop (Fig. 70).
• Insert a specimen and rotate a suitable objec-
tive into place.
• The current fluorescence cube is indicated on
the display.
• Using Booster Lens and Excitation Manager,
the Excitation Manager can be inserted into
the back receptacle.
• Closing the incident light shutter protects your
specimen from fading.
Do so by pressing the SHUTTER variable key.
(For key occupation please see “Identification
Sheet”.)
The display indicates the symbol: ↓
• Selecting the fluorescence filter cube:
Press the variable keys
Cube
or Cube
|
|
Fig. 69 Inserting the Booster Lens
Fig. 70 Inserting the Excitation Manager
57
Download from Www.Somanuals.com. All Manuals Search And Download.
9. Imaging Procedure for Leica DM4000 M
9. Imaging procedure
for Leica DM4000 M
9.1 Incident Light
9.1.1 Bright Field
9.1.2 Dark Field
• Switch to the incident light axis (IL) by pushing
the TL/IL button.
• Switch to the incident light axis (IL) by pushing
the TL/IL button.
• Select the DF (dark field) contrast method.
Do so by pressing the DF variable key.
• Select the BF (bright field) contrast method.
Do so by pressing the BF variable key.
Alternatively: Press the CHANGE RL |
variable key.
(For key occupation please see “Identification
Sheet”.)
Alternatively: Press the CHANGE RL
|
variable key.
(For key occupation please see “Identification
Sheet”.)
The display indicates DF.
The DF reflector is turned into the beam
path.
The display indicates BF.
• Insert a specimen.
• Insert a specimen.
• Rotate an appropriate objective into place.
• Rotate an appropriate objective into place.
• Bring the image into focus using the focus dial
and set the brightness using the INT function
key.
• Bring the image into focus using the focus dial
and set the brightness using the INT function
key.
Notes:
• The maximum objective aperture which may
be used for dark field is 0.75. All objectives
with greater aperture are automatically
blocked for this procedure ("DF" flashes in the
display).
• When selecting the dark field method, the
aperture diaphragm is opened completely and
may not be adjusted. To avoid errors in opera-
tion, the function keys for setting the aperture
diaphragm (AP) are locked.
58
Download from Www.Somanuals.com. All Manuals Search And Download.
9. Imaging Procedure for Leica DM4000 M
Automatic procedure:
9.1.3 Polarization
• Switch to the incident light axis (IL) by pushing • The ICR filter cube is automatically brought
the TL/IL button.
into the light path.
• Select the POL (polarization) contrast method.
Do so by pressing the POL variable key.
Alternatively: Press the CHANGE RL |
variable key.
(For key occupation please see “Identification
Sheet”.)
Mechanical procedure:
• Rotate the appropriate polarizer (71.3) and the
IC/P analyzer (72.1) on the stand manually into
the light path. Also bring the polarizer and
analyzer into cross position until they reach
maximum darkness.
The display indicates POL.
• Insert a specimen and rotate a suitable objec-
tive into place.
Fig. 71 Objective prism slide
1
2
3
Knurled wheel for fine focusing
Prism slot with inserted objective prism slide
Insert polarizer
Fig. 72
1
Insert analyzer
1
3
2
1
59
Download from Www.Somanuals.com. All Manuals Search And Download.
9. Imaging Procedure for Leica DM4000 M
9.1.4 Interference Contrast
9.2 Transmitted Light
• Switch to the incident light axis (IL) by pushing 9.2.1 Bright Field
the TL/IL button.
• Switch to the transmitted light axis by pushing
the TL/IL button.
• Insert a specimen and rotate a suitable objec-
tive into place.
• Select the BF (bright field) contrast method.
Do so by pressing the BF variable key.
• Select the DIC contrast method.
Do so by pressing the DIC variable key.
Alternatively: Press the CHANGE RL |
variable key.
(For key occupation please see “Identification
Sheet”.)
Alternatively: Press the CHANGE RL
|
variable key.
(For key occupation please see “Identification
Sheet”.)
The display indicates BF.
The display indicates ICR.
• Insert a transmitted light specimen.
• The ICR filter cube (containing polarizer and
analyzer) is automatically brought into the light
path on the incident light axis. Insert the objec-
tive prism slide into the prism slot (71.2).
• Rotate an appropriate objective into place.
• Use the focus dial to bring the image into fo-
cus and set the brightness using the INT func-
tion key.
Alternatively:
• Rotate the ICR polarizer (71.3) and the IC/P
analyzer (72.1) on the stand manually into the
light path.
• Insert the objective prism slide into the prism
slot (71.2).
• For fine adjustment, rotate the knurled screw
(71.1) on the objective prism slide.
60
Download from Www.Somanuals.com. All Manuals Search And Download.
10. Trouble Shooting
10. Trouble Shooting
Problem
Cause/Remedy
Stand
The microscope does not respond.
ꢄMake sure that voltage is impressed.
ꢄMake sure that the microscope is connected
to the power supply.
ꢄCheck the cable connections.
ꢄInform service technician to change the fuses.
Illumination
The image is completely dark.
ꢄOpen the shutter (→ p. 35).
ꢄCheck the connection of the lamp houses to
the microscope.
Transmitted axis:
Incident (Fluo) axis:
ꢀ
ꢀ
ꢄMake sure that the lamps are connected to the
power supply.
ꢄInform service technician to change the fuses
of the ebq 100.
The image is unevenly or not uniformly illumi- ꢄRemove all unneeded filters from the light
nated.
path.
ꢄCenter the lamp (→ p. 41ff).
ꢄReplace the old lamp (→ p. 20ff).
The illumination "flickers."
ꢄBe sure that there is no loose connection at
the power supply.
ꢄReplace the old lamp (→ p. 20ff).
The lamp does not illuminate immediately upon ꢄThe ebq 100 must be switched-on repeatedly.
being switched on.
ꢄHot Hg lamps should cool down before
switching on again.
61
Download from Www.Somanuals.com. All Manuals Search And Download.
10. Trouble Shooting
Problem
Cause/Remedy
Bright Field
The specimen can not be brought into focus.
ꢄUse the correct immersion medium.
ꢄLay the specimen with the cover glass to-
wards the top.
ꢄMake sure that the cover glass thickness is
correct and that is conform to the indication
on the objective.
ꢄCheck the condenser centering.
Dark Field
No definite DF contrast is possible.
ꢄBe sure that a DF objective is being used.
ꢄThe objective aperture setting is too high
(maximum 0.75). If necessary, reduce the ob-
jective aperture using the iris diaphragm on
the objective.
ꢄCheck the condenser centering.
ꢄThe magnification is too weak. Use a higher
magnification.
The image is unevenly or not uniformly illumi-
nated.
ꢄClean the specimen and neighboring lenses
Undesirable stray light
(→ p. 65).
Phase contrast
No phase contrast is possible.
ꢄThe specimen is too thick.
ꢄThe refraction index of embedding material
and object is identical.
ꢄThe cover glass is not placed evenly.
ꢄTherefore, check the centering of the light
rings (→ p. 39).
62
Download from Www.Somanuals.com. All Manuals Search And Download.
10. Trouble Shooting
Problem
Cause/Remedy
Polarization
No polarization contrast is possible.
ꢄBring the polarizer and analyzer into cross po-
sition until they reach maximum darkness
(without specimen) (→ p. 55, 59).
Fluorescence
The image is completely dark (no fluorescence). ꢄOpen the shutter (→ p. 57).
ꢄSelect the incident light axis (IL) (→ p. 36).
ꢄCheck the antigen-antibody combination.
The fluorescence is too weak.
ꢄInsert the Booster Lens (→ p. 29).
ꢄCenter the lamp (→ p. 41ff).
ꢄInsert a new lamp (→ p. 20f).
Display
The display flashes.
ꢄRotate an appropriate objective for the con-
trast method into the light path.
FAIL! appears.
ꢄCheck insertion of objectives, cubes, etc.
63
Download from Www.Somanuals.com. All Manuals Search And Download.
11. Care of the Microscope
11. Care of the Microscope
Caution!
Caution:
Unplug the power supply before performing
cleaning and maintenance work!
Protect electrical components from moisture!
Residual fiber and dust can create unwanted
background fluorescence.
!
Cleaning Coated Parts
Microscopes in warm and warm-damp climatic Dust and loose dirt particles can be removed
zones require special care in order to prevent with a soft brush or lint-free cotton cloth.
fungus contamination.
The microscope should be cleaned after each Clinging dirt can be cleaned with all
use, and the microscope optics should be kept commercially available water solutions, benzine
strictly clean.
or alcohol.
For cleaning coated parts, use a linen or leather
cloth that is moistened with one of these sub-
stances.
11.1 Dust Cover
Note:
Caution:
Acetone, xylene or nitro-containing thinner
can harm the microscope and thus may not
be used.
!
To protect against dust, cover the microscope and
accessories with the dust cover after each use.
Test cleaning solutions of unknown composition
first on a less visible area of the unit. Be sure
that coated or plastic surfaces do not become
matted or etched.
Caution!
Let lamps cool down before covering the
stand with a dust cover. The dust cover is
not heat-resistant. In addition condensation
water may occur.
Cleaning the Stage
Remove light-colored spots on the stage by rub-
bing with paraffin oil or acid-free Vaseline.
11.2 Cleaning
64
Download from Www.Somanuals.com. All Manuals Search And Download.
11. Care of the Microscope
Removing Immersion Oil
Cleaning Glass Surfaces
Remove dust on glass surfaces with a fine, dry
and fat-free hair brush, by blowing with a blow
bag or vacuum suction.
Caution!
Follow safety instructions for immersion oil!
Carefully remove stubborn dirt on glass surfaces
with a clean cloth moistened with distilled
water. If the dirt still can not be removed, use
pure alcohol, chloroform or benzine.
First, wipe off the immersion oil with a clean cot-
ton cloth, and then re-wipe the surface several
times with ethyl alcohol.
11.3 Handling Acids and Bases
Cleaning Objectives
Caution!
For examinations using acids or other aggres-
sive chemicals, particular caution must be
taken.
The objective may not be unscrewed during
cleaning. If damage appears on inner sur-
faces, the objectives must be sent to your
Leica subsidiary for repair. We also advise
against cleaning the inside surfaces of the
eyepieces.
!
Caution:
Be absolutely certain to prevent the optics
and mechanical parts from coming these
chemicals.
The front lenses of objectives are cleaned as de-
scribed under "Cleaning Glass Surfaces". The
upper lens is cleaned by being blown off with a
pneumatic pump.
65
Download from Www.Somanuals.com. All Manuals Search And Download.
12. Wear and Spare Parts
12. Essential Wear and Spare Parts
Order No.
Material No.
Name
Used for
Replacement Lamp
500 974
500 137
500 138
500 321
Halogen lamp
12 V 100 W
107/2 lamp housing
106 z lamp housing
106 z lamp housing
106 z lamp housing
High-pressure mercury burner 50 W
High-pressure mercury burner 100 W
High-pressure mercury burner 100 W
(103 W/2)
500 139
High-pressure xenon burner
75 W
106 z lamp housing
Objective turret
Screw cap for unused objective receptacles
020-422.570-000 Screw cap M 25
Replacement eyecup (diaphragm protection) for HC PLAN eyepiece
021-500.017-005
021-264.520-018
021-264.520-018
HC PLAN eyecup
HC PLAN eyecup
HC PLAN eyecup
10x/25 eyepiece
10x/22 eyepiece
10x/20 eyepiece
Immersion Oil conforming to DIN/ISO standards, fluorescence-free
513 787
513 522
513 788
110 ml
100 ml
500 ml
OIL and IMM objectives
and oil condenser heads
66
Download from Www.Somanuals.com. All Manuals Search And Download.
13. Abbreviations and Pictograms
13. Abbreviations and Pictograms
Contrast method
+
Magnification
ꢂ ꢃ
Light intensity/diaphragms
Beam splitting
Transmitted light shutter open
Transmitted light shutter closed
Incident light shutter open
Incident light shutter closed
↑
↑
↓
↓
AET
AP
Advanced Ergo Tube
Aperture diaphragm
Brifght field
BF
COMBI
CUBE
DF
Combination contrast method
Filter cube
Darkfield
DIC
FD
Differentiated interference contrast
Field diaphragm
FLUO
ICR
ICT
Fluo axis (incident light)
Interference contrast reflected light
Interference contrast transmitted light
Incident light (axis)
IL
INT
Intensity
MBDT
PH
Motorized Basic Documentation Tube
Phase contrast
POL
TL
Polarization
Transmitted light (axis)
67
Download from Www.Somanuals.com. All Manuals Search And Download.
14. Index
14. Index
Adjusting the light sources 41
Allowable ambient conditions 15
Ambient temperature 8, 9
Ambient conditions 15
Analyzer 28, 55, 56, 59
Analyzer cube 55, 56
Fluorescence 57
Fluorescence intensity 57
Focus finder 33
Focusing 37, 47
Focusing telescope 39
Function keys 12, 31, 36
Objective aperture 54
Objective prism slide 60
Objectives 19, 50
Objective turret 10, 12
Object stage 17, 46
Aperture diaphragm 12, 35, 38, 52
Phase contrast 39, 53
Phase contrast rings (checking) 39
Polarization 55, 59
Polarizer 55, 56, 59
Polarizer L/ICR 27
Gas discharge lamps 23, 24
Beam splitting 49
Bright field 53, 58
Booster Lens 29, 57
Halogen lamp 21, 22, 42
Hg 50 burner 24
Polarizer R/ICR 27
Cleaning 64
Cleaning objectives 65
Coaxial pinion 46
Condenser connector 18
Condenser 11, 12, 18
Condenser centering 38
Condenser height adjuster 17, 18
Condenser holder 18
ICT/P transmitted light polarizator 27
Identification Sheet 19, 26, 28, 53
Imaging procedure10, 53, 58
Immersion oil 50, 65
Polarizer R/P 27
Prism slot 60
Reflector cube 26
Reflector cube for lamp adjustment 41
Reset function 31
Retention pin 26
Rotating polarizer 27
Incident light axis 10
Incident light nosepiece disc 26
Incident light polarizators 27
Incident light shutter 35
Connection to power supply 30
Initialization 34
Interference contrast 60
Intermediate systems 16
Shutter 57
Software Leica DMControl 11, 14
Specimen holder 17
Dark field 54, 58
Dark field stop 54
Defined function keys 36
DICprisms 28
Köhler illumination 18, 37
Supply unit ebq 100 9, 25, 46
Switch transmitted/incident light 12
Differentiated interference contrast 56
Display 12, 35
DMControl 11, 14
Lamp housing 13
Lamp housing 106 z 21, 41
Lamp housing 107/2 20, 41
Touchscreen 11
Transmitted light and
Lamp housing receptacle 20, 22, 25 incident light analyzer 28
Electrical safety 8
Electronics box Leica CTR5000 8, 30
Ergomodule 29
Excitation Manager 10, 29, 57
Eyepiece 12, 49
Light intensity 35
Light sources 41, 52
Light sources (incident light axis) 21
Light sources (transmitted light axis) 20
Transmitted light axis 10
Transmitted light filter 13
Transmitted light shutter 35
Variable function keys 12, 13, 31, 36
Magnification changer 11, 51
Vision problems 49
Field diaphragm 12, 35, 52
Filter block exchanger 13
Filter cube 26
Mercury lamp Hg 50 W 43
Mercury lamp Hg 100W /Xe 75W 44 Xe 75 burner 24
Mirror housing 29
Filter cube ICR 59, 60
Motorized analyzer 28
FIM (Fluo Intensity Manager) 10, 52
Motorized polarizer 27
68
Download from Www.Somanuals.com. All Manuals Search And Download.
15. EU Declaration of Conformity
15. EU Declaration of Conformity
69
Download from Www.Somanuals.com. All Manuals Search And Download.
|