Haier Freezer DW 86L288 User Manual

User Manual For Haier DW-86L288/388/628/386/626  
DW-86W420  
Ultra Low Temperature Storage Freezer  
This product is suitable for ultra low temperature storage of products in  
applications such as clinical, pharmaceutical, scientific research, and  
epidemic institutions.  
Haier Medical and Laboratory Co., Ltd.  
Read this user's manual carefully before using the ULT freezer.  
Place this manual in a convenient location for future reference.  
0070505464  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Technical Service  
Features of the Haier ULT Freezer  
Your Haier ULT Freezer has a 3 year warranty on the compressors and 5 years parts  
and labor on the rest of the unit.  
The Haier ULT Freezer is equipped with microprocessor control and a  
digital temperature display with a range of -10  
operation temperature range is -50 to -86  
to -86 . The normal  
The warranty card should be kept for future service work. Should you need service or  
to inquire about your freezer, please use the warranty card to contact your local Haier  
Equipment and Instrument Service center.  
.
The unit has a number of standard protection functions. They include  
compressor delay protection, low voltage booster, and high voltage  
protection.  
The refrigeration system for the unit features a balanced cascade  
technology powered by low-temperature rated speciality compressor.  
Advanced LED display on the freezer can display the cabinet  
temperature, set point, ambient temperature and line voltage. It also  
allows the user to set the high temperature alarm, low temperature alarm  
and notifies the user that the unit has developed a failure through the  
display.  
Innovative and integrated design of the door lock and handle provide  
convenient application.  
Note: Technical data published in this manual are nominal. The actual  
ULT performance may vary slightly due to continuous product  
improvements.  
24  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
Optional Accessories  
Description of Key Parts and Functions  
Diagram of parts  
CO2 and LN2 backup cooling system  
For installation and instructions for operation, please refer to the user's manual shipped with the  
system.  
Control  
panel  
Warning Any equipment using CO2 can displace oxygen in the surrounding area. Please  
verify that there is sufficient ventilation around the workplace. If there is limited ventilation,  
alternative means to ensure a safer environment must be employed. This may include air  
quality monitoring and warning devices.  
Door  
handle  
Inner  
door  
Outer door  
Side panel  
for  
refrigeration  
compartment  
Storage rack and box  
If the unit is used to store small samples, storage racks and boxes provide more efficient use of  
internal space.  
Front protection  
panel for  
Storage box  
refrigeration  
compartment  
Racks  
Reserved  
window for  
temperature  
recorder  
Storage box  
Door  
handle  
Control  
panel  
Lock  
Model  
Box  
Storage Rack  
20 (2 inch)  
16 (2 inch)  
12 (2 inch)  
16 (2 inch)  
20 (2 inch)  
DW-86L628  
DW-86L388  
320 (2 inch)  
192 (2 inch)  
144 (2 inch)  
192 (2 inch)  
320 (2 inch)  
DW-86L288  
DW-86L386  
DW-86L626  
Protection panel  
for refrigeration  
compartment  
Power switch for battery  
23  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
Optional Accessories  
Safety Precautions  
This manual and labels attached to the product include important safety precautions  
required to operate this product. Customers are advised to follow instructions  
provided to avoid potential damage to product and injury to operators.  
Note: Use this product only in the way described in the product literature and this  
manual. Before using it, verify that this product is suitable for its intended use.  
Do not modify system components or use unauthorized parts as this will void the  
product warranty.  
This recorder has been calibrated at the factory. It retains the temperature accuracy  
and calibration even when there is a power supply interruption. Should there be a need  
to recalibrate the recorder, please follow the procedures below.  
a)Let the unit run continuously until it reaches a stable condition. The recorder has  
registered temperature for at least two hours.  
b)Place a standard thermocouple in a 100 ml container filled with 10% glycerin solution.  
Place the container and the temperature recorder sensor in the unit. Do not place the  
recorder sensor in contact with any liquid.  
This sign is directly related to operator safety and equipment safety. The  
contents and instructions that follow the sign must be strictly followed.  
c)After four hours of soak time at the stable freezer temperature, the container  
temperature should match the freezer temperature. Then compare the thermocouple  
reading with the temperature recorder reading. If there is a discrepancy, use the button  
(#1) or the button (#2) on the recorder to move the pen on the chart to reflect the  
correct temperature.  
This sign refers to prohibited actions. Violation may result in damage to the  
freezer or compromise the operator's safety.  
Note: The ink pen will operate 5 seconds after the button is released.  
This label is adhered to a cover that protects a high voltage area to  
reduce the chance of electric shock. Access to the cover should only be  
handled by a qualified engineer or technician.  
Caution: The temperature recorder is preset during production, so resetting of the  
temperature recorder is not recommended unless absolutely necessary.  
NOTICE!  
1.Freezer must be placed at least 200mm away from the wall to allow maximum air  
circulation.  
2.Changing of the Refrigerants should be done according to the Service Manual.  
NOTICE!  
For the most efficient energy consumption and optimized compressor life,the  
temperature should be set at the highest temperature suitable for the samples to be  
stored,normally set at -50 to -80  
.
When power to the unit is on, switch on the battery charger. When the unit  
is not in use switch power off and then switch off the battery charger..  
Grounding mark  
Wear Cryogloves before opening the door.  
22  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Optional Accessories  
Proper Usage  
The appliance requires 220V~/50Hz for 628/388/288/386/310 or 220V~/50Hz and  
220V~/60Hz for 628/626.A regulator with a capacity of 4000W or higher must be  
installed if the voltage if beyond the rated voltage of 10%.If the power cord needs  
to be extended, the extended conductor must have a cross area of equal or greater  
than 2mm2 for 628/388/288/386/310(220V~50/60Hz) or 3mm2 for  
628/626(115V~/60HZ ),and the extended length must not exceed 3m.  
2) Remove the supplementary power supply  
Power LED light continues to flash. The recorder works normally. It suggests that the  
main power supply is sufficient while the backup battery lacks capacity. It is then  
necessary to replace the battery until the green LED lights up and stops flashing.  
3) Changing of Recorder Chart Paper  
Your Haier ULT unit is equipped with a standard three-prong power plug, rated  
c u r r e n t f o r 2 0 A  
When changing the chart paper for the recorder, please use the following steps.  
a. Locate the button (#3) at the front of the recorder.  
f o r 6 2 8 / 6 2 6 ( 11 5 V ~ 6 0 H z ) a n d 1 6 A f o r  
628/388/288/386/626/310/420(220V~50/60Hz). Be aware that one of the three  
prongs is for grounding purposes. Removal of the ground prong is strictly  
prohibited for safety reasons under any circumstances. The electrical power plug  
should be securely plugged into the power socket. A loose plug in the socket may  
cause fire.  
b. Press and hold the button (#3) for about one second. The imprinting stylus of the  
recorder moves to the left side of the chart.  
c. When the stylus is completely off the chart, unscrew the center nut for the paper  
and remove the old chart paper off the recorder. Place a new chart paper on the  
recorder. Carefully match the day and time line on the chart with the reference mark on  
the recorder panel (there is a small groove on the front of the recorder as shown in the  
figure).  
The power socket intended for your Haier ULT usage must be grounded to avoid  
electric shock. If the socket does not meet this requirement, the condition must be  
corrected by a qualified technician before using the ULT unit.  
d. Place the center nut over the chart paper and tighten it. Press the button (#3) to  
move the stylus back to the chart to continue the temperature recording.  
e. Inspect the contact between the pen and the paper chart. Adjust the arm if  
necessary to make contact. (Note, do not damage the pen and arm. It might be easier  
to remove the arm and pen to make a small bending adjustment and then install the  
pen back on the recorder. Repeat the process until the recorder arm and pen work  
properly.)  
Never use gas lines, water mains, telephone lines or lightening rods as the  
grounding device for your Haier ULT unit. This type of improper grounding may  
cause electric shock or other danger.  
Never install your ULT in an unprotected area. After a unit is rained on, there is a  
danger of electric shock.  
f. The accuracy of the temperature must be checked after changing the chart paper.  
The procedure is described as follows. Press and hold the button (#3) until the pen  
moves to the outside of the chart paper. Then press the button (#3) again; the pen  
should return to the chart paper area. When the pen reaches the outer most circle of  
the chart paper, it pauses for a moment. If the pen does not match the outer circle of  
the chart marks, you can use the button (#2) or the button (#1) to move the pen to  
match the line. This must be done within 5 seconds after the pen pauses. If the  
calibration is incomplete, repeat the same process until the pen matches the outer  
circle line.  
Installation and maintanence of your Haier ULT should be carried out only after  
reading this manual. Haier Equipment & Instrument Services can install and  
validate your unit please call local Haier sales and service center.  
Your Haier ULT must be securely installed on a firm floor. A tilted installation may  
result in equipment falling causing injury and damage.  
Your Haier ULT must not be installed in a damp area or an area that can be subject  
to water spray. A damp or wet unit may cause weakening of insulation which could  
lead to electric shock.  
4) Calibration of Temperature Recorder  
Never store flammable, explosive or volatile materials in the unit. To avoid potential  
explosion or fire, do not use any spray near the unit.  
Never store corrosive chemicals with acid or alkaline properties in the unit as they  
can lead to damage to internal components of the unit.  
Do not use any hard object to explore any openings or gaps such as air ventilation  
ports. Accidental contact between the hard object and a moving part may result in  
electric shock or injury.  
If poisonous, radioactive or other harmful materials need to be stored in the unit,  
the equipment should be located in a safe zone. Improper usage of the equipment  
with this type of materials may hurt the environment or operator's health.  
21  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Optional Accessories  
Proper Usage  
1.Temperature recorder  
The ULT Freezer is equipped with a 6-inch temperature recorder for the chamber  
temperature. When the unit is powered up, the recorder starts working.  
For proper operation of the temperature recorder, please follow the procedures below.  
1.Open the recorder door to access the recorder.  
If there is leakage of petroleum gas or another flammable gas, close the gas supply  
valve and open doors and windows to ventilate the air. Do not plug or unplug your  
freezer unit.  
2.Plug in the 9 volt battery located on the upper right corner of the recorder. This  
battery is a standby power.  
3.Install a new chart by following the diagram below.  
While the equipment is repaired, take precautions not to inhale any chemicals or  
aerosols existing near the unit. They might be harmful to your health.  
Never directly pour water into the unit. The water may cause electric shock or  
short circuit.  
9V battery  
Button  
Customers must not dismantle, repair or modify the equipment. Such improper  
operations may result in fire or personal injury.  
Should there be a malfunction in the equipment, power off the unit and unplug the  
power cord from power supply. Continuous operation with an abnormal condition  
may result in electric shock and fire.  
Hand  
Dateline  
Button  
In the event there is a power outage and recovery, be sure to wait for 5 minutes  
before the unit power is turned back on. This is to reduce the chance of damage to  
the compressors and refrigeration system. During a repair operation, power must  
be disconnected. The power cord should always be protected to avoid damages.  
Screw & fixing  
Recorder  
When unplugging the power cord from the socket, use a firm grip on the plug itself  
for pulling. Do not pull the power cord as this action may strip the wires off the plug  
causing electric shock, electrical short circuit and fire.  
(Figure 5)  
The Temperature Recorder  
2.Power supply  
Maximum loading on the each shelf should be not more than 50kg and total loading  
for whole unit should be not more than 200 kg.Heavier loads may cause damage to  
the shelving system .  
The temperature recorder requires an AC power supply. If AC power supply is  
interrupted, the LED of the recorder blinks and the warning indicator turns on. The  
temperature recorder can also be operated with DC from spare source battery which  
lasts about 30 hours.(Caution: The battery should remain fully charged. If the unit is not  
in operation, the battery should be switched off. Ensure that the battery is in good  
status as a spare power source.) If power fails, the LED of the temperature recorder  
blinks until AC power is supplied or spare power source is connected. The LED  
indicator will blink when the power in the battery is low.  
Replacement of Recorder Battery  
When the green LED on the recorder flashes, there are two ways to inspect the unit.  
1) Supplementary power supply (9 volt DC)  
If the LED light is off and the recorder stops working, it would suggest that there might  
be issues associated with the main power supply. Please inspect whether there is a  
loss of power, a short circuit, or a loose connection. If there is no issue with the main  
power supply and the recorder starts working normally, please install the backup  
battery to the recorder. At this time, if the LED stops flashing, the supplementary  
power supply is normal. If the LED continues to flashes, it means that the power supply  
is under capacity requirement. It is necessary to replace the battery. The LED green  
light is off when there is sufficient power to the recorder.  
20  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Proper Usage  
Schematic Diagrams, Refrigeration and Wiring Diagram  
Do not touch any electrical components or power cord if your hands are damp or  
wet. Such action might lead to an electric shock.  
Refrigeration diagram  
After restarting your unit after a power outage or shutdown, ensure that all settings  
are correct. Accidental change in setting may damage the stored products.  
Filter dryer  
Capillary tube  
Filter dryer  
Evaporator  
Do not place any water container or heavy object on top of the unit. A falling object  
may injure an operator. If the water spills into the unit, it may damage the insulation  
causing electric shock.  
Cascade condenser,  
Capillary tube  
Do not climb on top of the unit or place any object on top of the unit. Falling  
equipment may cause injury or property damage.  
Condenser  
Suction  
Hold firmly onto the handle to close the door to avoid pinching your hands.  
Accumulator  
Low stage compressor  
High stage compressor  
Pressure switch  
Do not place any glass container or enclosed container in the freezer. These  
containers may crack at cold temperatures causing injury to operators.  
Do not use bare hands to directly handle the stored products. The cold temperature  
of the products and the interior wall may cause the skin to burn.  
Oil separator  
During any repair operation, gloves should be worn to avoid injury by sharp edges  
or corners.  
wiring diagram  
The air filter for the condenser should be cleaned regularly. Freezer temperature  
could be compromised.  
If a unit is out of service for a long period of time, make sure the power cord is  
unplugged. Deteriorated insulation on the power cord may lead to electric shock or  
fire.  
When a CO2 backup system is activated, the ventilation in the freezer room must be  
monitored. Increased CO2 concentration in the air can be harmful and even fatal.  
Do not tilt the unit more than 45 degrees when moving the unit.  
Note: The way of connection the transformer pin: for 120V: connect 1 with 5, connect 2 with 6; for 240V: connect 2 with 5.  
19  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Pre-Installation Information  
Description  
Model Description  
Ultra Low Temperature Storage Freezer  
DW-86L628/388/288/386/626 DW-86W310/420  
When a ULT unit has been placed in storage or out of service for a long time, its  
battery should be tested for low capacity because the battery may have already  
released all of its energy. It does not mean the unit is in a malfunctioning condition.  
Should this condition occur, please run the unit for about 6 days to fully charge up the  
battery.  
Exterior wall material Coated cold rolled steel  
Interior wall material Coated zine cold rolled steel  
Inner doors  
Shelves  
Aluminum framed PS board  
Adjustable stainless shelves  
Porthole for testing  
Insulation  
25.8 mm diameter X 1 Located on back wall  
When the unit operates normally, the unit frame at the front near the door may be  
slightly warm. This phenomenon is normal because hot tubing is embedded there to  
prevent condensation from forming on the frame.  
Vacuum insulated with polyurethane foam(non-CFC)  
Compressors  
High stage/Low stage: hermetically sealed  
Before samples are loaded into the unit, make sure that the unit temperature has  
reached the set point then load the samples into the freezer in batches. This method  
can reduce the temperature rise as the load is being introduced.  
Evaporator  
Condenser  
Refrigerant  
Copper tube  
High stage: Finned coil Low stage: tube in tube heat exchanger  
R134a R404a R290 R23  
The temperature display indicates the temperature where the temperature sensor is  
mounted inside the unit chamber. Temperature uniformity within the chamber may vary  
from the sensor location due to loading patterns or insufficient temperature stabilization  
time.  
Temperature controller Microprocessor controller  
Temperature display Digital display  
Temperature sensor  
RTD Pt 100 ohms  
High temperature alarm, low temperature alarm, power outage alarm, ambient  
temperature alarm, dirty condenser alarm, battery alarm, voltage alarm  
An access port is installed in the back wall of the unit which can be used as the  
through hole for the thermocouple wires during testing and validation . After all test  
wires are let through the access port, make sure that the gap in the port is sealed  
properly with insulation materials. Otherwise, the unit temperature may not come down  
to the desired temperature. The port ring in the outer wall can also accumulate frost  
and ice.  
Alarm device  
Remote alarm  
terminals  
Maximum load is 24 VDC at 1A  
Rechargeable 12 VDC, automatically rechargeable  
Defrost scraper  
Battery  
Accessory  
Note: design parameters may vary. No further notice is available.  
Technical data  
When cleaning the unit, mild or neutral detergent solution should be used. (Non-  
diluted cleaning chemicals can damage the plastic portion of the unit. For directions on  
making a cleaning solution, please refer to the detergent section in the instructions).  
Never use a hard wire brush, acid, gasoline, detergent powder, polishing powder, or hot  
water to clean the freezer as these tools and materials can damage the powder coating  
and surface of the plastic. Particularly, never use gasoline or a solution with volatile  
chemicals to clean plastic or rubber parts.  
L
After the freezer runs for some time, a layer of frost usually forms on the interior  
liner and inner doors. When this layer of frost gets too thick, it can negatively impact  
the refrigeration performance of the unit. Energy consumption can increase. If the  
thickness reaches about 5 millimeters, please use the provided scraper to remove the  
frost.  
Before removing the frost, temporarily transfer the stored samples to another  
freezer. This reduces the chance of property loss due to rising temperatures.  
Behind the interior walls, there are many refrigeration tubes. Do not use a knife, an  
ice pick, or a screwdriver to cut ice and frost. This can potentially damage not only the  
liner but also the refrigeration tubes.  
Note: At 25 ambient, the unit's maximum cold temperature can reach -86  
Packing list  
.
If the freezer is not in use for a long time, please turn off the power and switch off  
the backup battery. The power cord should be unplugged.  
Key  
/
/
/
2
2
/
18  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Recycle Rechargeable Battery  
Installation  
The Haier freezer is equipped with a rechargeable battery. This battery is recyclable.  
When the battery reaches the end of its life, please contact a local recycling  
organization for inspection or properly discard the battery.  
Environment  
1.Ambient temperature: The Haier ULT is designed to operate between 5  
.The maximum ambient temperature is 32 For best results, the operating  
ambient should be in between 18 to 25 If necessary, use an air-conditioning  
system to achieve the required ambient condition.  
to 28  
1.Location of the battery  
.
.
The battery in the freezer is for the power outage alarm. It is located inside the control  
box in the lower left behind the panel.  
2.Relative humidity in environment: At a normal temperature, the maximum relative  
humidity should be less than 80 percent. At an environment of 32 , the maximum  
relative humidity is 57 percent.  
3.The intended location should be low in dust count.  
4.The intended location should be vibration and shock free.  
Warning  
There are high voltage components in the control box. To prevent  
electric shock, a qualified technician or engineer should open the cover.  
5.The highest elevation the unit can work safely in is 2,000 meters from sea level.  
.
2.Removal of the battery  
6.Input voltage should be within 220 or 120 VAC 10  
Turn off the power to the unit and unplug the power cord from the socket.  
Use a screwdriver to remove the screws on the side panel and take down the side  
panel.  
Notes: 1. An ultra low temperature freezer is usually sensitive to its operating  
environment. If a unit is installed outside the conditions mentioned above, it cannot  
operate reliably. Please improve the environmental conditions before using the  
equipment.  
Use the screwdriver to remove the cover for the control box.  
Remove the bracket that fastens the battery. Remove the battery.  
Follow regulations to recycle the battery or discard it properly.  
2. It is prohibited to use the unit in an outdoor place. After the unit is rained on, there  
is a danger of electric shock.  
Caution  
Location for installation  
When changing the battery, you must make sure that the brown wire connects to  
the positive of the battery, and the blue wire connects to the negative of the battery.  
The polarity must not be reversed. Incorrect polarity can damage the charge board  
so that it cannot charge the battery.  
For the equipment in order to achieve an optimal operation condition, an intended  
installation location should satisfy the following requirements.  
1.Do not install the unit in a room where there is little room to work on the unit should  
there be a malfunction. The doorway should be large enough for the unit to freely  
enter or exit the room if necessary. This is to allow the unit be repaired easily to avoid  
damage to property.  
2.There should be at least 30 centimeters space all around the unit for good ventilation  
for heat removal.  
3.The location for installation should be flat and firm.  
Brown wire for the positive  
terminal of the battery  
4.There should be good ventilation and no direct sunlight.  
5.The freezer unit cannot share the same power socket with another piece of  
equipment. The power plug should be securely connected with the power socket.  
6.The power cord for the freezer should not be twisted or pinched.  
7.If there is a need to extend the power cord, the extension should be rated for 220  
VAC at 16 A and 120 VAC at 20A .  
8.Before using the freezer, check the voltage supply. A voltage stabilizer to deliver 220  
or 120 VAC 10 percent is recommended for areas where the voltage is known to be  
unstable. The voltage stabilizer should be rated at least 4 kilowatts.  
9.A proper grounding terminal must be guaranteed.  
Blue wire for the negative  
terminal of the battery  
10.If the power socket is connected with a ground terminal, make sure to inspect it for  
proper connection before using the equipment with it.  
11.If the socket does not have a ground connection, it must be installed by a qualified  
technician.  
17  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Initial Loading  
Troubleshooting Guide  
Installation  
Should there be any malfunctions in the system, please attempt to answer the  
following questions before notifying maintenance or calling Haier Equipment &  
Instrument Service Center  
1.Remove all packing materials and straps.  
2.Adjust support legs  
Rotate the leveling legs clockwise to extend them out to anchor to the floor to ensure  
the unit does not move while in usage.  
Freezer does not start up.  
Is the power supply normal? Has the main power switch been turned on yet?  
Is the voltage supply too low?  
Is a larger amount of load placed into the unit at one time?  
Leveling  
leg  
Leveling  
leg  
Freezer's refrigeration ability is reduced.  
Is the environment temperature too high?  
Are the inner doors and outer door closed properly? Does any ice and frost damage  
the seal between the door and the frame?  
Is the condenser filter clogged?  
3.Grounding  
Is the temperature setting correct?  
Is the freezer far away from any direct sunshine?  
Is the freezer near any heat source?  
Warning  
A socket with proper grounding must be used for the equipment. If  
the socket is not grounded, an upgrade of the power socket to include grounding  
must be done by a qualified technician. Do not use gas lines, water mains, telephone  
lines or lightening rods as the grounding device as they can lead to electric shock.  
After the unit's power cord is plugged into the power socket, the plug must be easily  
accessed. This allows a quick disconnect of the unit from the power supply in case of  
an emergency. The condenser inlet must not be blocked at anytime.  
Is the porthole plug installed in the porthole with proper insulation materials?  
Has the freezer been loaded with too many non-frozen samples within the last few  
hours? If so, the problem is temporary. You can set the temperature to a higher  
value to let the freezer cycle at temperature then gradually lower the set point  
temperature until it reaches the desired temperature.  
Test and adjustment  
The unit is noisy.  
Is the unit set on a firm floor?  
When the unit is started the first time, please follow the procedures below.  
1.Leave the unit standing in its intended location for at least 24 hours before  
switching the power on.  
Does the exterior of the unit touch any objects?  
Is the freezer unit leveled with the leveling legs?  
2.While keeping the unit empty, plug in the power cord to a dedicated power socket  
that meets all requirements.  
Does the freezer operate in a low temperature zone? If the load is heavy, and the  
cabinet temperature is warm, the unit may emit noise. The noise level would  
decrease as the cabinet temperature gets colder.  
3.Please turn on the power switch for the rechargeable battery located on the  
control box in the back the unit. This control box is visible from the back of the unit. If  
this switch is in the "OFF" position, there will be a low battery alarm during functional  
test.  
Is the noise emitted from the compressors? ULT compressor noise tends to be  
higher than that of a household refrigerator unit.  
The noise might be a crackling sound. This is probably the result of contraction of  
internal mechanical parts due to rapid cooling inside the equipment. Stepwise  
pulldown in temperature with slower temperature change can reduce this noise.  
4.If the unit has a backup cooling system, turn off the backup system.  
5.If the alarm buzzer sounds, press the cancel button to silent the alarm.  
6.Set the unit to desired temperature:Do not load the unit with any samples. Power  
up the unit to let it run down to -60 . Let it run at -60 for 8 hours then lower it to -  
80  
working properly.  
.
Observe the unit performance for 24 hours for normal cycling to ensure it is  
7.Once the unit is confirmed to operate properly, it is ready to be loaded with  
samples. In principle, the freezer unit should be set at about 3 above the desired set  
point. For example, if the storage temperature is -60 , set the unit at -57 . Load the  
unit to 1/3 of the unit's capacity. Make the unit is capable of cycling for more than 8  
hours.  
8.If the unit has a backup cooling system, turn on its control.  
16  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Initial Loading  
Routine Maintenance  
Warning:To prevent electric shock or injury to operators, the AC power supply to the freezer  
must be disconnected completely before any repair and maintenance work is to be performed.  
During any repair maintenance work, do not breathe in medical particles or aerosols near the  
equipment as they might be harmful to your health.  
Warning  
Your Haier ULT unit should be cared for by dedicated personnel. Every 2 to 4  
hours, the unit's working status should be inspected and recorded on a daily basis.  
Should there be a malfunction in the unit, the freezer temperature rises. If the  
problems cannot be corrected in a short time, please remove the stored samples;  
move them to another unit that meets the temperature requirement to avoid  
potential damage of property.  
Before loading the samples into the unit, make sure to check if the required  
storage temperature matches the temperature capability of the freezer. This is to  
avoid any property damage due to insufficient storage temperature for the  
samples intended to be stored in the freezer.  
Cleaning the freezer  
Clean the unit once a month. This can help the exterior look new.  
Use a dry cloth to wipe off loose dust inside and outside of the freezer. If the unit is rather dirty,  
use a clean cloth soaked with neutral detergent to clean the unit. Then use a dry cloth to wipe  
off the left over detergent solution. Non-diluted detergent can harm plastic surfaces. Please  
follow the instructions on how to mix a detergent solution.  
Do not pour water directly onto the exterior and interior of the unit. Water can damage the  
electrical insulation causing malfunction.  
Compressors and other mechanical parts are hermetically sealed. They do not need  
lubrication.  
Clean the frost and ice off the unit once a month. Also, clean the condenser filter once a month.  
Because it takes time for refrigeration temperature to reach the stored samples,  
there is normally  
temperature and the set point. This is a normal phenomenon. The lower the set  
point is, the smaller this temperature discrepancy becomes.  
a
temperature discrepancy between the actual shown  
Clean the condenser filter  
When the control panel shows an alarm signal for Hot Condenser, the alarm light flashes. The  
condenser filter must be cleaned to avoid compressor overheating and to improve refrigeration  
efficiency. Even if the light is not on, the filter should still be  
Filter  
cleaned once a month. To clean the filter, follow the procedure  
screen.  
below.  
Caution  
1.Pull off the front grill cover.  
2.Pull out the filter screen.  
3.Use water to wash the filter screen.  
All ultra low temperature storage units are low temperature storage equipment. It  
is prohibited to load an excessive amount of samples into the unit at one time.  
The compressors run for a long period of time without stopping. The freezer  
temperature may not decrease, and the compressors can become overheated.  
Samples must be loaded in batches, and in stepwise decreasing temperature  
setting. The process should be repeated until the final temperature is reached.  
Do not use any unauthorized mechanical means to remove or defrost the unit.  
Do not damage the refrigeration circuit.  
Front protection  
panel for refrigeration  
compartment  
4.Install the screen back in its original position and close the cover.  
5.If the Hot Condenser light is on before cleaning, check the light to make sure the light is off after  
cleaning.  
Defrost the interior  
Frost usually forms in the upper portion of the unit and the on the inner doors. Frost and ice can  
form in between the door gasket and frame to form an air gap. It can decrease the refrigeration  
effect of the unit. Follow the procedure below to defrost the interior doors using the plastic scraper.  
The first step is to move the samples to another freezer before the procedure. Do not use any  
sharp tools such as knife or screw driver to defrost.  
Do not use any non manufacturer-approved electrical components in the freezer.  
1.Remove the samples from the unit that needs to be defrosted. Move them to another unit or a  
container with a CO2 backup system for temporary storage.  
2.Turn off any backup refrigeration system if there is one.  
Operation after a power outage  
Your Haier ULT freezer control setting is stored in its memory system. Should there be  
a power outage and the power recovers, the unit can resume its operation based on the  
prior set control settings.  
3.Turn off the power supply.  
4.Open the outer door and inner doors to let the unit thaw for a period of time.  
5.Use a dry cloth to soak up and remove the water on the floor of the unit.  
6.After defrosting the unit and cleaning up the water, restart the unit.  
7.Load the samples back into the unit after it reaches the set temperature.  
8.Turn on the backup refrigeration system if necessary.  
Changing battery  
The battery that supports the power outage alarm is a consumable item. The life expectancy for  
the battery is about 3 years. If the battery is more than 3 years old, the battery should be replaced  
because the alarm function may not work properly , and the stored capacity could also be reduced.  
To do so, please contact Haier Equipment and Instrument Service center.  
10  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Temperature Adjustment and Setting  
Display and Alarms  
Notes:  
A flashing alarm light cannot be cancelled unless the root cause for the alarm is eliminated.  
The buzzing alarm can be temporarily silenced for 30 minutes by pressing the alarm silence  
button. However, if the problem is not fixed, the buzzer alarm will resume after 30 minutes.  
The freezer is equipped with a rechargeable battery controlled by a power switch. Before  
operating the freezer, this switch must be turned on. Once the unit is plugged into the  
specified voltage, the unit operates to charge the battery. When there is a power outage, the  
battery sustains the temperature display. If the battery voltage is less than a threshold, the  
temperature display turns off. While the battery is still capable of providing power to the  
display, the temperature display can be turned off by unplugging the power cord and turning  
off the battery control switch.  
Alarms  
Silence  
-80  
Run  
SET  
Voltage Regulator  
V
ULT FREEZER  
The procedure to change the freezer temperature is as follows.  
The freezer is also designed to auto-adjust the inner temperature set point at high ambient.  
When the ambient temperature is warmer than 32 and the set point temperature is set to be  
lower than -82 , the set point temperature will automatically default to -82 . If the ambient  
temperature is equal to or cooler than 32 , the set point will resume to the intended inner set  
point. This feature extends the life expectancy of the freezer.  
1.In order to change the setting of the freezer, the freezer's control must be unlocked  
first. The temperature display flashes when the up or down arrow is pushed. Push the  
up or down arrow again to change the display to security code mode then enter the  
initial value 06. Push the SET key for 5 seconds. The locking sign turns off. This  
means the unit is in unlocked mode. Push the SET key again to select the mode for  
the temperature set point value, high temperature alarm value and low temperature  
alarm value. The matching indicator lights up as each function is chosen.  
Buzzer alarm resumption after cancellation  
When the unit is in the alarm mode, you may press the Silence key to stop the buzzing of  
the alarm. However, the remote alarm cannot be cancelled. If the alarm condition still persists  
after 30 minutes, the buzzer alarm resumes automatically.  
2.Freezer temperature set point: Push the SET key. Once the Inner Temp.indicator  
is on, the temperature set point area flashes. To change the temperature set point,  
push the  
or  
key intermittently to increase or decrease the temperature  
Remote alarm terminals  
setting by 1 . After the adjustment is done, do not make any further changes for 10  
seconds. The temperature display stops flashing. The unit enters the locked mode.  
Remote alarm terminals are located in the machine compartment in the back of the unit. The  
alarm signal is delivered via the terminals. The terminals are rated for 24 VDC at 1 A. The power  
for remote alarm system must be connected with the wire or battery with UL certificate.  
The temperature range is -10 to -86  
.
3.Setting of high temperature alarm:Push the  
Alarm light. The temperature display flashes. To change the high temperature alarm of  
the unit, push the or key intermittently to increase or decrease the  
SET  
key to turn on the High Temp.  
Terminal output  
Under normal operating conditions, the terminals with wires are open. When the unit is in the  
alarm mode, the two terminals with wires are in the closed mode. The following schematic is a  
wiring diagram.  
temperature setting by 1 . After the adjustment is complete, do not make any further  
changes for 10 seconds. The temperature display stops flashing. The unit enters the  
locked mode. When the freezer temperature is warmer than the high temperature  
alarm set point, the unit's alarm sounds, and the high temperature alert light flashes.  
High temperature alarm set point can only be set to a value of at least 5 warmer than  
the set point temperature of the unit. The high temperature alarm value cannot be set  
to exceed the maximum limit.  
Remote alarm terminal with brown wire  
Remote alarm terminal with blue wire  
Alarm device  
DC power source  
Connect with remote  
alarm terminal with brown  
wire  
4.Setting of low temperature alarm: Push the SET key to turn on the Low Temp.  
Alarm light.To change the high temperature alarm of the unit,push the  
key intermittently to increase or decrease the temperature setting by 1  
or  
Connect with remote alarm  
terminal with blue wire  
.
After the  
adjustment is complete, do not make any further changes for 10 seconds. The  
temperature display stops flashing. The unit enters the locked mode. The low  
14  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Display and Alarms  
Temperature Adjustment and Setting  
Display and alarm  
When the freezer is powered up, the display will show the actual freezer temperature,  
set point temperature, and current voltage.  
temperature alarm set point can only be set to a value of at least 5 colder than the  
set point temperature of the unit. It cannot be set to the lowest temperature limit.  
The display shows the following indicator lights.  
Locked indicator:When the light is on, it means all set values are in the locked mode.  
This is to prevent erroneous operations.  
5.Setting the security code: When the unit is first setup to operate, the security  
code to unlock the control is 06. After unlocking the unit, simultaneously push SET  
Network indicator:The indicator suggests that the network is in working mode.  
Run indicator:When the light is on, it means the compressors are in working mode.  
Regulator indicator:The lit indicator means the voltage stabilizer is in working mode to  
boost or buck the voltage.  
and Silence for 5 seconds. The display will show 06. Use the  
and  
key to change the security code. The choices are from 05 to 30. After a number is  
chosen, do not touch the unit for 5 seconds. The control system will enter the locked  
mode. The new code is effective.  
Backup System indicator:The lit indicator means that the backup cooling system is in  
working mode.  
6.Setting the Startup delay timer: In order to reduce the startup power surge after a  
power outage, your Haier ULT incorporates a startup delay feature. This feature can  
alter the startup sequence of the compressors in the freezer. After the control is  
Alarm Conditions  
Alarm  
Alarm Trigging Condition  
Indicator  
Buzzer Sound  
unlocked, simultaneously push  
SET key and  
key for 5 seconds. The  
High  
When this indicator is on,  
the units temperature  
reaches the warm alarm limit.  
Buzzer delays 15  
minutes to sound  
intermittently.  
Buzzer delays 15  
minutes to sound  
intermittently.  
temperature display shows the time delay to be 01 for 1 minute for the high stage  
compressor to start after power up. This delay can be changed by pushing the  
Alarm indicator flashes.  
Temperature  
Alarm  
Low  
Temperature  
Alarm  
When this indicator is on,  
the units temperature  
reaches the low alarm limit.  
or  
key to change the value from 1 minute to 10 minutes. The default value for  
Alarm indicator flashes.  
this delay is 1 minute. If this time delay startup is set at more than 5 minutes, the  
equipment would take too much time to pull down the freezer temperature. If the power  
source has sufficient capacity, there is no need to change the default value.  
Power Failure  
Alarm  
Alarm indicator flashes.  
Display shows temperature  
and blank interchangeably.  
Buzzer delays for 1  
minute to sound  
intermittently.  
Equipment loses power.  
7.Alarm test and low battery test:  
When the ULT unit is powered up,  
Voltage Improper  
Alarm  
Buzzer delay for 1  
minute to sound  
intermittently.  
simultaneously press the "Up" arrow and "Alarm Cancel" buttons for 5 seconds. The  
buzzer alarm should sound and the alarm indicator should flash. At this condition, the 6  
second test procedure starts. If the battery capacity is low, the low battery light indicator  
will flash 6 times. If the battery is at full capacity, the low battery light does not light up  
or flash. After 6 seconds, start the unit up. All indicators will light up for 6 seconds  
without flashing. All electronic display windows will indicate a digit of 8 for 6 seconds.  
When this is done, all display functions work properly.  
Supply voltage  
exceeds the high or low limits.  
Alarm indicator flashes.  
Alarm indicator flashes.  
1.Condensers filter element  
is clogged.  
2.Ambient emperature is too  
high.  
Hot Condenser  
Alarm  
Buzzer sounds  
intermittently.  
When ambient  
High Ambient  
Temperature  
Alarm  
Ambient Temperature  
.
temperature exceeds  
38 , buzzer sounds  
intermittently.  
Alarm indicator flashes.  
exceeds 32  
Alarm indicator flashes, temperature  
display alternates to show E2 and  
cabinet temperature.  
Alarm indicator flashes, temperature  
display alternates to show E1 and  
cabinet temperature.  
Buzzer sounds  
intermittently.  
Main cabinet temperature  
control sensor fails.  
Probe Failure  
Alarm  
Buzzer sounds  
intermittently.  
Condenser sensor fails.  
Ambient sensor fails.  
Alarm indicator flashes, temperature  
display alternates to show E0 and  
cabinet temperature  
Buzzer sounds  
intermittently.  
Alarm indicator flashes, temperature  
display alternates to show E3 and  
cabinet temperature.  
Buzzer sounds  
intermittently.  
Heat exchanger sensor fails.  
Low battery  
Alarm  
Battery capacity runs low or  
battery switch is not turned on.  
During the test, the alarm  
indicator flashes.  
----  
12  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Graco High Chair 1761279 User Manual
Haier Freezer HCF150 User Manual
Honeywell Smoke Alarm 5808W3 User Manual
Hotpoint Freezer HMB312AAI User Manual
Hotpoint Freezer RC15P User Manual
HP Hewlett Packard Child Tracker 8444A User Manual
HP Hewlett Packard Server BL2X220C User Manual
Hypertec Carrying Case 5703 User Manual
Hyundai Appliance Trim Kit 95 7333 User Manual
iHome Stereo System IP39 User Manual