Haier Freezer DW 40L508 User Manual

Medical Device Manufacturing Enterprise License No.:  
No. 20050143 license granted by Shandong Food and Drug Administration for  
manufacturing medical devices  
Registration No.:  
No. 2580116 in 2008 approved by Shandong Food and Drug Administration for  
manufacturing medical devices (updated)  
This series of freezers is in conformity with the registration standard YZB/Lu  
0174-2009.  
Medical Low-Temperature Freezer  
Operating Instructions  
Sincere Forever  
Certificate of Quality  
Inspector:  
DW-40L508  
Haier Group  
[Company name]:  
[Registered address]:  
Qingdao Haier Medical and Laboratory Products Co., Ltd.  
Haier Industrial Park, Qianwangang Road, Economic and Technological  
Development Zone, Qingdao  
[Manufacturing address]:  
Haier Industrial Park, Qianwangang Road, Economic and Technological  
Development Zone, Qingdao  
0532-88936043  
0532-88936010  
266555  
It applies to the storage of various articles at low temperature by the clinical,  
medical, scientific research and inspection departments, etc.  
Please read this manual carefully before use.  
[Tel.]:  
[Fax]:  
[Postal code]:  
Special No.:0270500015  
Version:1st version in 2009  
Please keep this manual for future reference.  
It is only suitable for sale and use in Chinese Mainland.  
V98497  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Components  
Safety precautions  
Control  
panel  
Please always follow all important safety regulations in this manual; otherwise you  
must take full responsibility for the consequences.  
Please follow all precautions and regulations in this manual to avoid possible injury  
of the user and any other person.  
Handle  
“Caution”, “prohibition” and “warning” signs are as follows:  
Door snap  
close  
The contents marked with this “caution” sign involve safety of both the  
product and user and must be strictly followed.  
Outer door  
The contents marked with this “prohibition” sign describe absolutely  
prohibitory behaviors and must be strictly followed; otherwise the freezer  
may be damaged or the user may be injured.  
Front cover of  
compressor  
compartment  
This sign is marked on the covers of high-voltage electric appliances to  
avoid electric shock. These covers shall be removed by a qualified  
engineer or maintainer.  
Drawer  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Safety precautions  
Safety precautions  
If poisonous, harmful or radioactive substances are stored in it, please use the  
freezer in a safe area. Improper use may be harmful to health or environment.  
In case of leakage of inflammable gas such as coal gas, please close the valve  
involved in gas leakage, open windows for exhaust ventilation and do not plug in or  
unplug the freezer.  
Be sure to disconnect the freezer from its power supply before carrying out any  
repair or maintenance on the freezer in order to avoid electric shock or personal  
injury.  
Make sure that no drug or suspended particles in and around the freezer will be  
breathed in at the time of repair or maintenance; otherwise they may result in health  
hazard.  
Never pour water directly onto the freezer; otherwise electric shock or short  
circuit may occur.  
Please use special power supply (AC 220V~/50Hz) marked on the freezer’s  
rating plate. If the voltage used is below 198V or above 242V, an automatic voltage  
stabilizer of 4000W and above should be additionally installed. If power cord needs  
to be extended, the extension cord must not be less than 2mm2 in sectional area and  
more than 3m in length. Using the power supply of any voltage or frequency other  
than that marked on the rating plate may result in fire hazard or electric shock.  
The freezer’s power cord is provided with a three-pin (grounded) plug which is  
suitable for standard three-wire (grounded) socket. In any case, do not cut of remove  
the grounding pin of power cord. Be sure to ensure firm connection and good contact  
between the plug and socket; otherwise fire hazard may occur.  
Please use a power socket with ground wire to avoid electric shock. If the power  
socket is not grounded, be sure to ask a qualified technician to install grounding wire.  
Never ground the freezer through gas pipe, water pipe, telephone line or lighting  
arrester; otherwise it may result in electric shock or other dangers.  
The freezer must not be disassembled, repaired or modified by the user himself.  
If an unauthorized person does so, fire hazard or personal injury may occur due to  
improper operation.  
In case of abnormal operation, please unplug the freezer. If the freezer goes on  
operating under abnormal state, electric shock or fire hazard may occur.  
If the freezer is left unused for a long time in an unsupervised area, please keep  
children away from it and make sure that its doors cannot be fully closed.  
If the freezer is turned off, it should be turned on again after at least 5 minutes to  
avoid damage to the compressor or system. Be sure to unplug the freezer before  
carrying out any repair or maintenance. Never roll or damage the power cord.  
Equipment scrapping should be done by qualified persons. The doors should be  
removed to avoid accidents such as suffocation.  
Never use the freezer in the open air. If it is exposed to rain, electric leakage or  
electric shock may occur.  
The freezer must be installed only by a qualified technician or maintainer;  
otherwise electric shock or fire hazard may occur.  
The freezer must be installed firmly on solid ground. In case of unfirm ground  
or improper location, it may topple or cause personal injury.  
When disconnecting the freezer’s plug from the power socket, please grasp it and  
do not pull the power cord; otherwise electric shock or fire hazard due to short circuit  
may occur.  
Never touch any electrical component such as power plug or any switch with a  
wet hand; otherwise electric shock may occur.  
When turning it on again after power failure or power-off, be sure to check the  
freezer’s settings. The change of settings may deteriorate those articles stored in the  
freezer.  
Do not put any container filled with water or any heavy object on the freezer. If  
the container or heavy object drops, personal injury may occur; in addition, the water  
may reduce the insulation level, resulting in electric leakage or electric shock.  
Do not climb the freezer or put any article on the freezer; otherwise it may topple,  
resulting in personal injury or damage to the freezer.  
Never install the freezer in a humid place or in a location exposed to  
splash/spray water; otherwise electric leakage or electric shock may occur due to  
lower insulation level.  
Neither store inflammable, explosive or volatile substances in the freezer nor  
use flammable spray near the freezer; otherwise explosion or fire hazard may occur.  
Never store corrosive products such as acid and alkali in the freezer; otherwise  
internal components or electric parts may be damaged.  
Never insert any metal object such as iron nail or iron wire into any hole/gap or any  
vent for internal air circulation; otherwise the contact between these objects and  
moving parts may cause electric shock or personal injury.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Safety precautions  
Precautions for use  
If the freezer is turned on for the first time or has not been used for a long time, its  
rechargeable battery may at low charge level because it has been fully discharged.  
This is not a fault. In this case, make the freezer operate continuously for about 6  
days to fully charge the battery.  
When the freezer is operating, the surface around the doorframe of shell may get  
hot. This is not a fault. The reason is that anti-condensation pipes are installed around  
the opening to prevent condensate from forming around the cabinet shell.  
Make sure that the freezer’s internal temperature has been low enough before  
putting articles in the freezer and the articles are put in the freezer in batches to  
prevent temperature rise.  
When closing a door, please hold its handle to prevent your fingers from being  
pinched.  
Do not store glass bottles or canned products in the freezer because frost crack  
may occur and result in personal injury.  
Do not touch any item stored in the freezer with a bare hand. Touching a frozen  
item or the freezer’s inner wall with a bare hand may cause frostbite.  
At the time of repair, wear a pair of gloves to avoid injury by a sharp edge or  
corner.  
The displayed temperature of low-temperature freezer means the temperature at  
the temperature-sensing probe in the preservation compartment. Sometimes it is  
different from actual temperature in the center of preservation compartment, but it  
will be gradually close to the actual temperature.  
If the freezer is to be left unused for a long time, pull out its power plug to avoid  
electric shock, electric leakage or fire hazard resulting from the aging of power cord.  
Put plastic bags out of the children’s reach because they may cause suffocation.  
There is a through-hole on the freezer’s rear panel for leading out the test wire  
inside the freezer. After leading out the test wire, be sure to stuff the test hole with  
heat-insulating material again; otherwise the temperature inside the preservation  
compartment will not be low enough and condensate may form around the outside of  
through-hole.  
Clean the freezer with diluted neutral detergent (undiluted detergent will damage  
the plastic composition. Please refer to the detergent instructions for the dilution  
method). Do not use a brush, acid, gasoline, soap powder, polishing powder or hot  
water to clean the freezer because they may damage the painted surface and  
plastic/rubber parts. The point we must pay special attention to is that any volatile  
solvent such as gasoline should not be used on plastic/rubber parts.  
Precautions for handling  
After the freezer operates for a period of time, there will be a layer of frost on its  
inner wall. Thick layer of frost will affect the freezer’s refrigerating effect and increase  
power consumption. Therefore defrosting should be done by the provided scraper or the  
defrosting key on the control panel when the thickness of frost reaches about 5mm.  
Before defrosting, please take out all articles stored in the freezer and put them in  
another freezer for preservation to avoid the deterioration of articles due to  
temperature rise.  
There are many cooling coils on the shelves. Therefore do not remove the frost on  
the shelves with any sharp tool such as knife, ice chisel or screwdriver, etc. Be  
careful not to scratch the evaporating pipes in the process of defrosting; otherwise the  
freezer will be unable to work normally.  
1. Before unpacking, a fork lift or bale clamping device may be used to move the  
freezer.  
If a fork lift is used to move the freezer, insert its fork into the bottom of wood  
bottom bracket from both sides of the freezer.  
If a bale clamping device is used to move the freezer, clamp the freezer from the  
bottom of wood bottom bracket and only clamp both sides of the freezer.  
2. After unpacking, the freezer can be moved by its castors.  
If the freezer is to be left unused for a long time, please disconnect it from the  
power supply and turn off the battery switch.  
Note: Never tilt the freezer more than 45° when moving it.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation  
Use  
Installation  
Installation environment  
1. Ambient temperature: 5~28; maximum temperature 32; ideal temperature  
18~25; if necessary, please use the air conditioning system.  
2. Ambient humidity: less than 80% RH. If maximum working temperature is 32,  
the ambient humidity should be less than 57% RH.  
1. Remove all packing materials and tapes  
Remove all packing materials and tapes for transport.  
2. Adjust the legs  
3. Avoid a lot of dust.  
4. Avoid mechanical swing or vibration.  
5. Altitude of the working position: less than 2000m.  
6. Input voltage: 220V~±10%.  
Clockwise turn the adjustable legs to extend them out and rest on the floor. Make  
sure that the freezer is not moved when it is in use.  
3. Grounding  
* Notes:  
Warning  
1. The low-temperature freezer is relatively sensitive to ambient temperature, and  
therefore it cannot operate normally if it is installed in any environment other than the  
above-mentioned. In this case, please change the environment and then use the freezer.  
2. It is prohibited to install the freezer in the open air. If the freezer is exposed to rain,  
electric leakage or electric shock may occur.  
Please use a power socket with ground wire. If the power socket is not grounded,  
be sure to ask a qualified technician to install proper ground wire.  
Never ground the freezer through gas pipe, water pipe, telephone line or lighting  
arrester, etc.; otherwise electric shock may occur.  
The power socket must be accessible after installation in order to disconnect the  
power cord from the power socket in an emergency.  
The freezer’s air vent must not be covered by anything.  
Installation location  
For normal operation and optimum performance, the freezer’s installation location  
should meet the following requirements:  
1. The freezer must not be installed in a narrow and enclosed space. The door body  
of the room where the freezer is installed must not be smaller and lower than the  
freezer and must ensure normal pass-through of the freezer to avoid difficulty in  
repair in case of any fault; otherwise the freezer cannot be restored in time,  
resulting in deterioration of the articles stored in the freezer.  
Commissioning  
When use the freezer for the first time, please follow the steps:  
1. After installation, be sure to leave the freezer still for at least 24 hours before  
turning it on.  
2. When the freezer is empty, connect the power cord to a special socket of proper  
specifications.  
2. At least 30cm should be reserved around the freezer for ventilation and heat dispersion.  
3. The freezer must be installed on flat floor.  
3. Turn on the rechargeable battery switch (it can be seen from the freezer’s back)  
on the electric cabinet of compressor compartment at the top of freezer’s back. If  
the switch is not turned on, low battery charge alarm will be given at the time of  
test.  
4. The freezer must be installed in a well-ventilated place without direct sunlight.  
5. The freezer should not share a socket with any other device. Ensure firm  
connection between the plug and the socket.  
6. Do not twist or press the power cord.  
4. When alarm sound is heard, please press the “Cancel buzz” key to stop buzz.  
5. Set the desired temperature: turn on the freezer without putting anything in the  
freezer and observe the freezer for normal start and stop at least 24 hours. If so,  
it means that the freezer’s performance is normal.  
7. If the power cord needs to be extended, the extension cord must be of 220V~ and 15A.  
8. Please check the working voltage before use and consider using a voltage  
stabilizer of 3kw and above in those areas with unstable voltage to ensure normal  
input voltage within 220V~±10%.  
9. The freezer must be grounded reliably.  
6. After confirming normal performance, put articles to be stored in the freezer.  
10. If the power socket is equipped with ground wire, please check it for reliable  
grounding before use.  
If the power socket is not equipped with ground wire, be sure to ask a qualified  
technician to install proper ground wire.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Use  
Temperature adjustment and setting  
The freezer’s temperature is set and  
adjusted as follows:  
Caution  
1. Unlocking  
The freezer should be in the charge of a specially assigned person. He/she checks  
and records the freezer’s running condition (every 2-4 hours). And, in case of a fault  
or power failure, the temperature inside the freezer will rise. If the fault cannot be  
cleared in a short time, please take out all articles stored in the freezer and put them  
in another freezer meeting their temperature requirement to avoid deterioration.  
Before putting articles to be stored in the freezer, make sure that the freezer’s  
temperature range meets the temperature requirement of these articles. Otherwise  
they may deteriorate because the freezer cannot reach the required temperature,  
resulting in loss.  
When the “Lock” indicator lights up, all keys except the “ ” key are locked.  
Press and hold the “ ” key for 5s, the display area for internal temperature will  
show F0 and “0” will flash; then press the “ ” key, the display area for internal  
temperature will show F1; finally press the “Set” key, the keyboard will unlock  
and the display panel will show internal temperature. If any key is pressed in 10s,  
the keyboard will automatically be locked.  
2. Temperature setting  
In initial power-on state, press the “Set” key, the display screen will show L0 and  
“0” will flash; press this key again, the display screen will show the default  
temperature setting (i.e. -30) and the second digit will flash. (If the freezer is  
not used for the first time, the display screen will show the temperature set last  
time) Adjust the freezer to the desired temperature by pressing the “ ” key to  
switch the flashing digit and pressing the “ ” key to change the value of current  
flashing digit, and then press the “Set” key to confirm and memorize the  
temperature setting. Temperature setting range: (-20~-45)  
Due to the inertia of refrigeration, there is a certain difference between actually  
displayed temperature and set temperature (the lower set temperature, the less  
temperature difference and vice versa). It is normal.  
3. High temperature alarm setting  
After unlocking, press the “Set” key, the display screen will show L0 and “0”  
will flash; press the “ ” key, the display screen will show L1; press the “Set” key  
again, the display screen will show the default temperature setting for high  
temperature alarm with the second digit flashing (the default temperature setting  
for high temperature alarm is internal temperature setting +10. Note: When  
showing the temperature, the “ ” sign followed by digits means the “+” sign)  
and then adjust the temperature setting by pressing the “ ” key to switch the  
flashing digit and pressing the “ ” key to change the value and sign of current  
flashing digit; after finishing setting, press the “Set” key to memorize the setting.  
Temperature setting range for high temperature alarm: maximum displayed  
temperature ≥ set temperature for high temperature alarm - set temperature ≥5;  
otherwise use the default setting.  
Prohibition  
All low-temperature freezers are storage equipment. It is strictly prohibited to put  
too many relatively hot items in the freezer at a time; otherwise the compressor will  
keep working for a long time without stop and be damaged easily and the  
temperature will not reduce. Be sure to put articles in batches and make the  
temperature inside the freezer reduce in steps to the desired temperature.  
Do not accelerate defrosting by using any tool or method not approved by the  
freezer’s manufacturer.  
4. Low temperature alarm setting  
After unlocking, the display screen will show L0 and “0” will flash. Press the  
“ ” key twice, the display screen will show flashing L2; press the “Set” key, the  
display screen will show the default temperature setting for low temperature  
alarm with the second digit flashing (the default temperature setting for low  
temperature alarm is internal temperature setting -10), and then adjust the  
Do not damage the refrigerating loop.  
Dot use any appliance not approved by the freezer’s manufacturer in the freezer.  
Operation after power failure  
The freezer can memorize the settings. In case of power failure, the freezer will still  
operate with the parameters set before power failure when power supply resumes.  
temperature setting by pressing the “  
” key to switch the flashing digit and  
pressing the “ ” key to change the value of current flashing digit; after finishing  
setting, press the “Set” key to memorize the setting. Temperature setting range  
for low temperature alarm: minimum displayed temperature ≤ set temperature for  
low temperature alarm - set temperature ≤-5; otherwise use the default setting.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Display and alarm  
Display and alarm  
Upon connecting to the power supply, the freezer will be on and the display screen  
will show the freezer’s internal temperature.  
Prompts for the working state of display panel:  
“Lock” indicator: if the indicator lights up, it means that all keys are locked to  
avoid misoperation.  
“Alarm” indicator: if the indicator lights up or flashes, it means that a fault occurs.  
Functions of keys:  
Cancel buzz: in the locked or unlocked state, press the “Cancel buzz” key at the  
time of alarm, the alarm sound will be cancelled and the visual alarm will not be  
cancelled.  
Alarm test: it is used to test normal working state of the alarm system and low  
battery charge level. In the unlocked state, press this key, all the indicators will light  
up, the buzzer will give an alarm and all digital tubes display “8”. In the locked state,  
press this key at the time of alarm, corresponding fault alarm code will display.  
Please identify the fault causes according to the table below.  
Note:  
Always turn on the battery switch before starting the freezer. The freezer will enter  
normal working state and charge the battery upon connecting to an AC power supply  
meeting its requirements. In case of AC power failure, the battery will energize the  
display screen for normal display. When the voltage of battery reduces to less than a  
specified value due to discharge, it will stop energizing the display screen and thus  
the display screen will not display any longer. When the battery is able to supply  
electricity normally, to completely cut off the freezer’s power supply, just pull out its  
AC power cord and turn off its battery switch. Then the display screen will not  
display any longer.  
Time setting for automatic alarm recovery  
At the time of alarm, press the “Cancel buzz” key on the display panel, buzz alarm  
will be cancelled (remote alarm will not be cancelled).  
If alarm conditions remain, the buzz alarm will automatically recover after a pause of  
30min.  
Alarm  
High  
State  
Fault  
code  
internal E9  
Indication  
Buzzer  
When  
the  
The  
indicator When the freezer is  
turned on for the first  
time, the buzzer alarm  
function is activated  
after 5 hours according  
temperature temperature is higher  
alarm  
flashes  
than the set temperature  
for high temperature  
alarm  
Remote alarm terminals  
The remote alarm terminals are located in the compressor compartment at the back of  
freezer. Alarm signals are sent by these terminals with carrying capacity of DC 30V, 2A.  
Contact output:  
In normal conditions, both ends of connecting wire of remote alarm terminals are  
open circuit. In alarm conditions, both ends of connecting wire of remote alarm  
terminals are closed circuit.  
To use the remote alarm function, connect the circuit as shown in the figure below.  
Note: The remote alarm function is limited to three alarm modes - high temperature  
alarm, low temperature alarm and power-down alarm.  
to the default settings; in  
normal operation, the  
buzz alarm is given after  
15min  
Low  
When  
the  
internal E10  
The  
flashes  
indicator  
temperature temperature is lower  
alarm  
than the set temperature  
for low temperature  
alarm  
Power-dow Power-down of the  
/
The  
alarm the buzz alarm is given  
n alarm  
freezer  
indicator  
flashes;  
screen  
after 30s  
black  
and  
actual internal  
temperature  
show alternately  
Over-high  
ambient  
Ambient  
above 38  
temperature E14  
The  
indicator  
flashes  
temperature  
Abnormal  
sensor  
Internal main sensor at E6  
fault  
The  
flashes  
The  
indicator Buzz alarm  
indicator  
Ambient  
temperature E0  
sensor at fault  
Insufficient  
charge or the battery  
switch is not on  
flashes  
The  
flashes 3 times  
Low  
battery E5  
indicator  
battery  
charge  
level  
Note: If several alarms sound simultaneously, each fault code displays 3s.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Routine maintenance  
Fault clearance  
Warning  
If the freezer goes wrong, please check the following possible fault points before  
asking for repair:  
Failure to start  
To avoid electric shock or personal injury, be sure to disconnect the freezer from power supply  
before carrying out any repair or maintenance.  
Make sure that no drug or suspended particles around the freezer will be breathed in at the  
time of maintenance; otherwise they may result in health hazard.  
Is the connection of power supply normal? Is the main power switch on?  
Is supply voltage too low?  
Were many articles put in the freezer at a time?  
Poor cooling  
Cleaning of the freezer  
Clean the freezer once a month. Regular cleaning makes the freezer always brand-new.  
Wipe off a little dust on the shell, inner compartments and all accessories of the freezer  
with a dry cloth. If the freezer is very dirty, remove the dirt with a clean cloth and neutral  
detergent, then wipe off the residual detergent with a wet cloth, and finally wipe the freezer  
with a dry cloth. (Undiluted detergent may damage plastic parts. Please refer to the detergent  
instructions for the dilution method.)  
Is ambient temperature too high?  
Are the doors closed tightly? (The frost between the cabinet body and the door  
seal may reduce the sealing performance of doors)  
Is the temperature setting correct?  
Never pour water onto the shell or into the preservation compartment of freezer; otherwise  
the insulation layer may be damaged, resulting in faults.  
Is the freezer away from direct sunlight?  
Is the freezer close to a heat source?  
Are the rubber hole cover and heat-insulating materials of test through-hole placed  
correctly?  
The compressor and other mechanical parts are fully enclosed and require no lubrication.  
Remove the frost or ice on the inner wall once a month.  
Were many uncooled items put in the freezer in past several hours? (In this case,  
the problem is temporary. You may raise the set temperature and then gradually  
reduce it after the freezer starts and stops normally in a short time.)  
Noise  
Removal of frost on the inner wall  
Frost usually forms on internal shelve, which may cause clearance between the cabinet body  
and door seal and thus result in poor refrigerating effect. You may use the ice scraper provided  
along with the freezer for defrosting. The following procedure is the method of defrosting  
through emptying the freezer.  
Is the freezer installed on solid ground?  
Is the freezer’s shell contact with any object?  
Note: Do not use any tool with sharp edges or corners such as knife or screwdriver for  
defrosting.  
1. Take out all articles stored in the freezer and put them in another freezer or in a container  
with freezing liquid carbon dioxide.  
Has the freezer been leveled by adjusting its legs?  
Is the freezer in low-temperature running state?  
If the freezer’s internal temperature is relatively high due to many items in it,  
sometime it make noises. The noise will lower with the reduction of temperature in  
the preservation compartment.  
Please check whether the noise is from the compressor (the compressor noise of  
low-temperature equipment is louder than that of household appliances)  
Knocking sound. It may be caused by rapid cooling and contraction of internal  
components. In this case, please make the freezer’s temperature reduce gradually  
without abrupt temperature reduction in a short time.  
2. Turn off the power switch.  
3. Open outer doors and leave them open for a period of time for defrosting.  
4. Wipe off water at the bottom of cabinet body with a dry cloth.  
5. After cleaning the cabinet body, turn on the freezer again.  
6. Put back the articles in the sufficiently cooled freezer.  
Or you may use the “Defrost” key to defrost: after unlocking the keyboard, press and hold the  
“Defrost” key for 5s, the compressor will stop running. This key doesn’t have automatic return  
function. Therefore it is necessary to press and hold it for 5s again after defrosting ends, and  
the compressor will resume normal running.  
Battery maintenance  
When the low-temperature freezer keeps working continuously, please test the battery  
charge level every 15 days (please refer to the section of alarm test function on Page 12 for the  
test method). In case of low battery charge level, make sure that the battery switch is on and  
the battery is charged. When the battery has been charged for consecutive 7 days, please retest  
the battery charge level. In normal conditions, the battery charge should be full. If low battery  
charge level remains, it is recommended to replace the rechargeable battery.  
The battery with power-down alarm function is a consumable item with 3-year service life.  
If the battery has been used for more than 3 years, it may not act at the time of alarm. It is  
recommended to replace the battery earlier. Please contact a Haier after-sales service  
technician for the battery replacement.  
Equipment scrapping  
Warning  
If the freezer is left unused for a long time in an unsupervised area, please keep  
children away from it and make sure that its doors cannot be fully closed. Equipment  
scrapping should be done by qualified persons. Be sure to remove the doors to avoid  
accidents such as suffocation.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Rechargeable battery recycling  
Specifications  
Technical data  
Model  
Ambi  
ent  
tempe  
rature  
()  
Classif  
ication  
of  
electric ()  
shock-  
Tempe Effe  
Rated  
Rat  
Inp Wei Mixe  
Overall  
dimensi  
Conn  
ection  
type  
A rechargeable battery is built in the freezer and is recyclable. When the battery’s  
service life expires, please contact local battery recycling organizations for check or  
proper disposal of the battery.  
rature  
range  
ctive voltage/fr ed  
volu  
me  
ut  
ght  
d
equency  
cur  
rent we  
(A)  
po  
(Kg refrig ons  
)
erant  
(g)  
(D×W×  
H)(mm)  
(L)  
r
1. Position of the battery  
proof  
protect  
ion  
(W  
)
The battery built-in the freezer is used for power-down alarm and located in the  
electric cabinet.  
DW-4  
0L508  
10~32  
I
-20~-4 490  
0
220V~/5  
0Hz  
4.2  
63  
0
170 280  
840×89  
5×1860  
Y
Caution  
Note: With ambient temperature at 25, minimum temperature inside the freezer can  
reach -40.  
There are high-voltage electrical components in the electric cabinet.  
To avoid electric shock, the cover of electric cabinet must be opened only by a  
qualified engineer or maintainer.  
Name  
Qty.  
Warranty  
Card  
Operating  
Instructions  
Plastic bag  
Ice scraper  
Thrust shelf Door handle  
subassembly  
Model  
DW-40L508  
2. Removal of the battery  
1
1
1
1
2
2
1) Turn off the power switch and pull out the plug from the power socket.  
2) Screw off the screws on the cover of electric cabinet with a screwdriver.  
3) Pull out the battery’s connecting plug.  
Note: Thrust shelves are installed at the back of freezer to ensure a certain distance  
between the freezer and wall.  
4) Remove the battery’s fixing clip and then take out the battery.  
5) Recycle or properly dispose the battery as required.  
Refrigeration principle and circuit diagram  
Schematic refrigeration diagram  
Prohibition  
When replacing the battery, make sure that the brown wire is connected to the  
battery’s positive pole and the blue wire to the battery’s negative pole. Do not  
misconnect; otherwise the charging circuit of computer board would be burned out  
easily and the battery cannot be charged.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Refrigeration principle and circuit diagram  
Options  
Installation and running  
In order to operate and use the recorder correctly, please operate as follows:  
1) You will see the recorder upon opening its door.  
2) Switch on the backup power supply, i.e. 9V DC battery in the upper left corner of recorder.  
3) Change a piece of new recording paper (see the figure below).  
4) Remove the plastic cap from the recording pen and close the recorder’s door.  
Note: The recorder will not work until internal temperature reaches the recorder’s recording range.  
Options  
1. Temperature recorder  
Ask a qualified technician or an after-sales maintainer to install the temperature recorder as follows:  
1) Remove the cover plate of temperature recorder.  
2) Install the temperature recorder in place and then fix it to the front panel with screws.  
3) Remove the screws of electric cabinet in the compressor compartment and then connect the  
connecting wires of temperature recorder respectively to the freezer’s live wire and neutral  
wire.  
4) Lead the sensor of temperature recorder in the freezer through internal sensor duct at the back  
of compressor compartment. The sensor duct is from the sensor through-hole at the back of  
freezer to the compressor compartment.  
5) Disassemble the holder of internal temperature sensor in the cabinet, fix the recorder’s sensor  
and internal temperature sensor together, seal the sensor hole on the cabinet body and install  
the cover in place.  
6) Bundle the connecting wires of recorder in the compressor compartment with cable clips.  
7) Fix and restore the cover of electric cabinet and the left panel of compressor compartment  
respectively with screws.  
Power supply  
When the freezer works, the recorder uses AC power supply in normal conditions. If AC power  
supply goes wrong, the recorder’s LED indicator will flash to indicate abnormal power supply. And,  
the recorder will go on recording internal temperature by using backup battery. Each backup battery  
can support the recorder for about 30h (Note: In case of low battery charge level, replace the battery  
in time to avoid corrosion of battery switch due to overdue replacement. To save battery charge,  
please turn off the battery switch when the freezer is not in use. Please install a backup battery in  
place to ensure normal recording in case of power failure during normal use). The recorder’s LED  
indicator will continuously flash until main power supply (AC power supply) is switched on and the  
backup battery is replaced. In case of low backup battery charge level, the recorder’s LED indicator  
will flash to indicate the need of battery replacement.  
Warning  
Repair and replacement of the recorder’s backup battery  
Before installing the temperature recorder, be sure to switch off the power supply; otherwise  
electric shock or fire hazard may occur.  
If the recorder’s green LED flashes, there are two checking methods:  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Options  
After-sales service  
The freezer is guaranteed for free by our company for one (1) year calculated from the date of  
purchase (the invoice shall prevail). Main components (including compressor, condenser, evaporator,  
main control panel and display panel) are guaranteed for free by our company for three (3) years.  
Warranty card: when you need common consultation or repair, please contact our local  
after-sales service center with both warranty card and purchase invoice on hand and keep the  
1. Remove the backup battery (9V DC), if the LED goes out and the recorder stops working, it  
means that the recorder’s main power supply goes wrong. In this case, please check for power  
failure, short circuit and loose connecting wire, etc. If the main power supply is in good  
condition after repair, the recorder will work normally. Then install the backup battery in place,  
if the LED stops flashing, it means that the backup battery is normal; if the LED goes on  
flashing, it means that the backup battery is at low charge level and needs replacement until  
the green LED lights up and does not flash.  
2. Remove the backup battery, if the LED goes on flashing and the recorder works normally, it  
means that the main power supply is normal and the backup battery is at low charge level and  
needs replacement until the green LED lights up and does not flash.  
warranty card properly.  
Operation Steps and Precautions for the Low-temperature Freezer  
Unlocking  
When the “Lock” indicator lights up, all keys except the “  
locked. Press and hold the “ ” key for 5s, the display area for internal  
temperature will show F0 and “0” will flash; then press the “ ” key, the  
display area for internal temperature will show F1; finally press the “Set”  
key, the keyboard will unlock.  
” key are  
Beware  
electric  
shock!  
There  
such  
components  
of  
Change of recording paper  
Change the recording paper as follows:  
are  
Temperature setting  
In the unlocked state, press the “Set” key, the display screen will show L0  
and “0” will flash; press this key again, the display screen will show the  
former temperature setting with the second digit flashing. Adjust the  
1. Find out the “Press” button (3#) in the upper left corner on the recorder’s front panel.  
2. Press and hold this button (3#) for about 1s until the recording pen starts moving towards the  
left of recording paper.  
3. When the recording pen entirely moves out of the recording paper, loosen the center bolt and  
change the used recording paper with a piece of new recording paper by carefully aligning the  
time line on the paper with the time slot on the panel (there is a small slot on the left of panel.  
Please see the attached figure).  
as  
control  
freezer to the desired temperature by pressing the “  
” key to switch  
board  
and  
the flashing digit and pressing the “ ” key to change the value of current  
flashing digit, and then press the “Set” key to confirm and memorize the  
temperature setting. Temperature setting range: (-20~-40)  
Precautions  
contactor,  
etc. in the  
electric  
1.  
After initial installation or movement, please adjust the legs and  
make them rest on the floor to make sure that the freezer will not move  
during use, then leave the freezer still for at least 24h before turning it on.  
cabinet.  
4. Retighten the center bolt, then press and hold the button (3#) to make the recording pen reset  
and start recording temperature.  
2.  
When using the freezer, maximumly control the room  
5. Check whether the recording pen writes well on the paper or not. If not, you may adjust its  
beam to make the penpoint contact with the recording paper. (Note: Do not damage the  
penpoint and beam. If it is difficult to adjust, you may remove the beam with a screwdriver,  
bend it with a little force and reinstall it in place by repeating the above operations).  
6. To ensure the recorder’s accurate recording, please calibrate the recording pen after each paper  
change as follows: press and hold the button (3#) until the recording pen moves out of the  
recording paper, then press and hold the button (3#) again until the recording pen return to the  
recording paper, at which time the pen will pause at the outmost temperature graduation mark  
on the recording paper (this graduation mark is not necessarily with temperature value, but it  
must be the outmost graduation mark). If the pen does not pause in this position, you may  
adjust its position by using 1# or 2# arrow within 5s to make the penpoint align with the  
outmost graduation mark. If you don’t finishing the adjustment of pen within 5s, please repeat  
step 6.  
Label for protection against electric shock:  
Warn the user not to open the electric  
cabinet at random.  
temperature below 28 (If ambient temperature is above 32 , the  
cooling efficiency will reduce rapidly. Also, if ambient temperature is  
above 32 for a long time, the compressor will be damaged or its  
service life will reduce. In this case, it is recommended to install an air  
conditioner).  
3.  
carrying current above 16A and reliable grounding.  
4. Before power-on, please turn on the battery switch on the electric  
The freezer should use an independent power socket of the  
cabinet at the back of freezer. If the freezer is to be left unused for a long  
time, please power off first, and then turn off the battery switch.  
Before power-on, please turn on the  
battery switch on the electric cabinet at  
the back of freezer. If the freezer is to  
be left unused for a long time, please  
power off first, and then turn off the  
battery switch.  
5.  
Maximumly control the time of each opening within 1min, and  
wipe off all ice and water on the door seal before closing the freezer’s  
doors to ensure high sealing performance.  
6.  
To extend the freezer’s service life and reduce energy  
consumption, it is recommended to set the temperature in the range of -20  
~-40 on the premise that the safety of items in the freezer is  
guaranteed.  
Calibration and adjustment of the recorder  
7.  
The freezer is used for the storage of items at low temperature  
The recorder has been accurately adjusted before leaving the factory and will not be affected even if  
power interruption. If it indeed needs adjustment, please do as follows:  
1. Keep the freezer operating continuously and then make it record the temperature curve for at least  
2h after it enters stable state (the fluctuation of displayed temperature is relatively regular).  
2. Put a standard thermocouple in a bottle with about 100ml of 10% glycerite, and then put the  
bottle and the recorder’s sensor together in the freezer (the recorder’s sensor must not be  
contact with any solution).  
3. The glycerite bottle with thermocouple is stored in the freezer for about 4h. When the  
temperature of glycerite in the bottle is identical with the freezer’s internal temperature,  
compare temperature values of the standard thermocouple and recorder, if they are not the  
same, you may adjust the recorder’s recorded value by using the adjusting buttons on the left  
(1#) or right (2#) of panel to make it in conformity with the thermocouple’s indicated value.  
Label for the use of rechargeable battery:  
Describe the precautions for the use of  
rechargeable battery  
and not used for quick freezing. It must not be forced to quickly freeze a  
large quantity of hot products or large-volume liquid products.  
8.  
If the freezer does not refrigerate in 2-3 hours after power on,  
please unplug it and contact our after-sales service enter as soon as  
possible.  
Rechargeable battery switch  
9.  
In case of alarm or any other fault, please clear the fault  
according to the prompts on the display panel. If there is no prompt or  
you cannot clear the fault, please do not disassemble the freezer by  
yourself but immediately contact our after-sales service center to ask a  
qualified maintainer for fault clearance.  
Label for the rechargeable battery switch:  
Show the position of rechargeable battery  
switch  
Fault codes are as follows:  
Grounding sign  
High temperature  
alarm  
Low temperature  
alarm  
Over-high ambient  
temperature alarm  
E9  
Main sensor fault  
alarm  
Ambient temperature  
sensor fault alarm  
Low battery charge  
E6  
E0  
E5  
Label for operation instructions and  
precautions:  
Instruct the user in operation and relevant  
precautions  
E10  
E14  
Note: The recording pen starts moving 5s after the button is released.  
Note: The recorder has already been calibrated before leaving the factory; do not adjust it except  
for special cases.  
level  
or  
battery  
switch off  
______Haier Group customer service hotline (for the whole  
country): 4006-999999  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Graco Baby Swing 1494 User Manual
Graco Inc Power Roller 309322D User Manual
Groupe SEB USA T FAL Iron FV6050 User Manual
HannsG Computer Monitor HH181AP User Manual
Hasbro Robotics 80735 User Manual
Heat Glo LifeStyle Indoor Fireplace TWILIGHT II User Manual
Hotpoint Dishwasher HDA3430ZWW User Manual
HP Hewlett Packard Camcorder PW550 User Manual
Hunter Fan Outdoor Ceiling Fan TypeB User Manual
Imation Camcorder Hard Disk User Manual