Haier Freezer DW 150W200 User Manual

Super Low Temperature (SLT) Freezer  
DW-150W200)  
Operation Manual  
Contents  
Before Use  
Product Features-----------------------------------2  
Parts Name  
-----------------------------------3  
Safety Precautions----------------------------------4  
Installation Location-------------------------------6  
In Use  
Preparations and Operation Procedures------7  
Control Board Adjustment---------------------9  
Maintenance & Servicing  
Defrosting · Non-use-----------------------------12  
Cleaning--------------------------------------------13  
Q & A-----------------------------------------------14  
After-sales Service------------------------------- 15  
Technical Parameter  
Schematic Circuit Diagram------------------ 18  
Please read this manual carefully before using the freezer.  
Please keep this manual in a safe place for future reference.  
The freezer is designed for preserving various articles at low temperature for such departments as  
clinic, medicine, scientific research and quarantine, etc.  
1-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Parts Name  
Door closure  
Control board  
Compressor  
compartment  
Cabinet  
Front cover of compressor compartment  
Notice  
Dear Haier’s customers,  
Thank you for choosing and using Haier products!  
Haier is dedicated to providing high-grade, high-precision and advanced  
products, and thus it may say that Haier products are featured with good quality  
and high performance. For your better use, please read this manual carefully before  
use and follow the operation procedures as described in this manual.  
Our “international star-level service” will accompany with you all the time. Any  
problem in use, please contact us via the telephone number and address given on the  
warranty card. We are willing to serve you at any time.  
Thank you again for choosing Haier products and recommending them to other  
people!  
Haier – sincere forever!  
Note: The product is suitable for applications in special laboratories. Please see actual product for its  
appearance, color or pattern.  
3-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Safety Precautions  
Please read all the installation and operation instructions carefully, and strictly follow them when  
installing and operating, otherwise the user shall be solely liable for any resultant damages (including  
part damage).  
Precautions for installation location  
1. Installation site of the equipment should ensure a well-ventilated circumstance, avoid direct  
sunlight.  
2. The freezer must not be installed close to heat sources and no heater is permitted to close it.  
3. The freezer must not be installed in an enclosed or narrow space.  
4. The room temperature should be kept below 25(if above 30, the refrigeration efficiency would  
lower rapidly and the compressor would be damaged easily).  
5. To be installed, the freezer should be kept at least 30cm away from any wall.  
6. The site where the freezer is used must be provided with a special air switch with leakage protector of  
proper specification (3-phase 16A, single-phase 25A); the wiring must be in the form of crimp  
connection and any socket/plug is strictly forbidden to use for connection.  
7. In case of frequent occurrence of unstable voltage or below rated voltage, a voltage stabilizer should  
be installed.  
Precautions for operation  
1. After its installation or handling, the freezer must be left still at least 24 hours before being energized.  
2. The freezer must operate at initial temperature setting (-126) for over 24 hours.  
3. After the freezer operates for 24 hours stably at initial temperature setting, set the operating  
temperature at -135and have the freezer stably operate at this temperature for at least 7 hours.  
4. If the freezer works normally at the second temperature setting, set target operating temperature.  
5. Normal range of internal temperature: -126to -150.  
6. Set alarm temperature as required (recommended alarm temperature: operating temperature plus  
15~25).  
Both refrigerant and foaming agent used by the freezer are flammable and explosive substances.  
Therefore, be sure that the site where the freezer is used is well ventilated and fire fighting equipment is  
provided near the installation location.  
Do not put too many or too hot samples into the freezer at a time, otherwise result in the increase of  
system load, prolonged running of the freezer and further burn-out of compressor. It is advisable to put  
4-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
samples into the freezer in batches and pre-cool those hot samples fully before putting them into the  
freezer.  
In case of handling or moving, only after the freezer is left still for over 24 hours can it be energized  
to operate, otherwise result in serious damage to the freezer, even compressor burn-out.  
For any equipment may go wrong or fail to work, the user must ensure that reserve equipment or  
reserve devices are available in the vicinity so that samples can be moved into safe reserve equipment or  
system for preservation in an emergency.  
It is required to have special equipment in-progress checked (at least 2 hours/time) by a full-time  
technician so that protective measures would be taken immediately in case of any fault. The  
manufacturer shall not be liable for compensation for any loss incurred by samples damage resulting  
from the users failure to find a fault and move stored samples into reserve equipment in time.  
In the case of ambient temperature above 27, it is required to open doors and windows, or turn on  
the air-conditioning system; while in the case of ambient temperature above 32, it is forbidden to use  
the freezer, otherwise that may result in compressor burn-out.  
Warning  
1. Do not load the freezer with samples over 1/5 of its capacity at a time.  
2. Do not put large-volume liquid samples or hot articles into the freezer.  
3. Do not leave the door open for over 1 minute.  
4. Note: The freezer is not for quick freezing and production, but for super low temperature (SLT)  
reservation. Therefore, it must not be forcibly used for the quick freezing of a large amount of hotter  
(relative to -140) samples.  
Exclusions:  
Equipment failure or samples damage resulting from the user’s failure to follow the above-mentioned  
operation instructions is not covered by the scope of warranty and compensation.  
Equipment failure or samples damage resulting from such reasons as artificial destroy and  
misoperation of the user, etc. is not covered by the scope of warranty and compensation.  
Equipment failure or samples damage not resulting from the manufacturer’s fault, but from fire  
hazard, war and earthquake, etc. is not covered by the scope of warranty and compensation.  
5-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Installation Location  
1. The freezer must be installed in a well-ventilated place where the room temperature is below 27°C.  
2. The freezer must not be placed in the site that under direct sunlight or close to heat sources.  
3. The temperature of equipment working environment should not be less than 10 ° C.  
The freezer should not be used if ambient temperature below 10°C.  
4. If ambient temperature is controlled by an air-conditioner, be sure to keep the air-conditioner working  
around the clock (24h).  
For good ventilation, the freezer should be at least 30cm away from a wall or other objects; otherwise the  
heat produced by its compressor may not be dispersed in time, resulting in the reduction of condenser’s  
cooling effect, even compressor burn-out in case of overload.  
5. The supply voltage should be within the range of 380V±50Hz 10% (or 220V±50Hz 10%). In case of  
frequent occurrence of unstable voltage, a voltage stabilizer (7.5 kW or above) must be installed.  
6. The freezer should not be installed in a humid place or a place to which water is easily splashed. The  
splashed water and dirt on the freezer should be wiped off with a soft cloth. It is recommended to use  
fire prevention system or dehumidification system.  
6-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Preparations and Operation Procedures  
When handling, it is strictly forbidden to overturn (upside down) the freezer; never allow its angle of  
inclination over 45° to avoid oil spill onto the system pipeline.  
Note: If the freezer is excessively inclined or upside down, compressor oil would spill onto the  
system pipeline, resulting in inadequate compressor oil. Further, inadequate compressor oil leads to  
insufficient lubrication, which results in noise from compressor, even agglutination of piston and  
cylinder inner wall, damage to compressor motor.  
Preparations before use  
When unpacking, all the packing materials (including base) should be removed.  
All the packing materials removed should be kept in a safe place and not thrown away before  
completing the commissioning. Only when the freezer is accepted after commissioning can you throw  
them away or destroy them to avoid inconvenience in commissioning or fault handling.  
Unpack the freezer with even force to prevent damage to the freezer or personal injury.  
After unpacking, first clean the freezer as described in the Section “Care and Maintenance”.  
The freezer should be installed horizontally in a well-ventilated place where the room temperature is  
below 27. The castors should be leveled and locked; otherwise noise and vibration may occur.  
To ensure no noise and vibration during running, please check:  
1. Whether the ground of installation location is flat.  
2. Whether all the castors are leveled and closely contact with the ground.  
Please install 380V/50Hz (3-phase live wire, single-phase null line, single-phase earth wire) or  
220V/50Hz (single-phase live wire/null line/earth wire) power supply strictly following the requirements  
specified in the product’s label, and have qualified installation personnel carry out wiring. Note: the  
capacity of distribution system must be higher than the rated capacity of freezer and the freezer must be  
reliably grounded.  
It is strictly forbidden to connect earth wire and null line; otherwise the freezer may be electrified  
and cause personal injury.  
To prevent the freezer from being affected by voltage fluctuation, please install a voltage stabilizer,  
the capacity of which should be higher than the rated capacity of freezer.  
Commissioning  
After finishing installation, be sure to leave the freezer still for over 24 hours before turning it on in  
order to ensure complete oil return.  
If the freezer is moved, only after leaving it still for 24 hours can perform any operation.  
In case of handling, if the freezer is energized before reaching stable state, the following problems may  
occur:  
1. It causes the symptom of compressor being lack of oil, resulting in noise produced by metal  
friction or knock;  
2. It causes the symptom of agglutination between piston and cylinder inner wall, resulting in  
damage to compressor.  
Before turning on the freezer, check the power supply for input voltage/power 380V/50Hz or  
220V/50Hz with a test pencil or a voltmeter, and check the earth wire for reliable grounding.  
Turn on the freezer, cool down in the sequence of -126, -135and minimum (target) operating  
temperature as described in the Section “Temperature Setting”, and the commissioning is completed  
once it works normally at minimum operating temperature. It is required to record the time of  
commissioning and corresponding temperature in details in the process of test.  
After completing commissioning, the results must be mutually confirmed by commissioning  
engineers and the user, and then put on file.  
7-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Samples storage and instructions  
Only after the freezer operates stably after commissioning and making sure that internal  
temperature meets storage requirements can samples be put into the freezer in batches.  
The freezer must not be loaded with samples over 1/5 of its effective capacity at a time. If samples  
are too hot, they should be pre-cooled down to required temperature by other equipment before being put  
into the freezer. After loading the freezer with samples, set internal temperature 5higher than the  
temperature of samples, keep it running until 8 hours after normal starting/stopping, and then decrease  
the temperature setting by -5~-10; continue such a cycle until reaching required temperature, and 12  
hours later put the 2nd batch of samples into the freezer.  
The freezer is not used for quick freezing, but for storage and preservation. Therefore, do not try to  
freeze relatively hot samples, or that may result in damage to samples and the freezer. What’s worse, that  
may result in destroy of samples and compressor burn-out. Furthermore, it is advisable to put fewer  
samples into the freezer at a time.  
Note:  
1. Do not put all the samples into the freezer at a time, but put 1/5 each time at an interval of  
24 hours.  
2. Be sure to wear special gloves in handling samples to avoid frostbite.  
Preparations required by safe samples storage  
For safe samples storage, the following preparations must be made to deal with emergencies such as  
power supply failure and equipment damage:  
1. Install a backup LN (liquid nitrogen) system.  
2. Provide a LN tank (can be preserved for 3~4 hours).  
Safe samples storage  
1. In an emergency, immediately move the samples stored in the freezer to a safe place and keep them  
well. The manufacturer shall not be liable for any sample damage resulting from the user’s failure to  
follow the instructions at any time.  
2. Ensure that a standby super low temperature freezer or LN tank is available in the vicinity for  
emergencies. The possible emergencies include equipment damage or fault due to its defects, or  
equipment damage for external factors such as fire hazard, power failure, etc.  
8-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Control board Adjustment  
Functions and operating instructions of the SLT freezer control board  
Temperature display  
A large display screen indicates actual internal temperature.  
A small display screen indicates current input voltage (no voltage indication for 3-phase power). Another  
small display screen indicates current ambient temperature.  
Prompt for working state:  
“Locked” indicator light “on” indicates all the settings are locked to prevent misoperations.  
“Network” indicator light “on” indicates the network system is in working state.  
“Running” indicator light “on” indicates the compressor is running.  
Voltage stabilizing” indicator light “on” indicates the voltage stabilizer is boosting voltage.  
“Backup system” indicator light “on” indicates the backup system has been connected.  
Alarm indication:  
In case of any fault required alarm, alarm is carried out by flashing light alarm signals and audible buzzer  
alarm. The former can not be cancelled until the fault clearance, while the latter can be silenced by  
pressing the “Cancel buzz” key.  
A. “High temperature” indicator light “on” as an alarm indicates internal temperature is higher than the  
high-temperature set point.  
B. “Low temperature” indicator light “on” as an alarm indicates internal temperature is lower than the  
low-temperature set point.  
C. “Voltage out of given range” indicator light “on” indicates the voltage is lower or higher than the lower  
limit or higher limit of normal voltage range.  
D. “Over-high ambient temperature” indicator light “on” indicates ambient temperature is above 30.  
E. “Sensor fault” indicator light “on” indicates a sensor fault.  
F. “Power failure” indicator light “on” indicates the power failure of power circuit.  
G. Dirty condenser” indicator light “on” indicates the condenser is blocked up by dirt and must be cleaned.  
9-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
H. “Low battery charge” indicator light “on” indicates the battery is at low charge and must be recharged;  
when connected to AC power supply, the SLT freezer charges the battery automatically.  
Key definition:  
There are 4 keys on the temperature control board. With the “Locked” indicator light on, all the keys are  
locked to prevent misoperations.  
“Function selection”: In unlock mode, press this key to select the desired function from “Internal temperature  
setting”, “High-temperature alarm setting” and “Low-temperature alarm setting” and the corresponding indicator  
light will be on.  
“Cancel buzz”: In lock mode, when an alarm occurs, press this key to silence the buzz for one hour, but the alarm  
by flashing light signals is not cancelled.  
The ” or “” key is for the increase or decrease of values.  
Starting and temperature setting  
The freezer’s control system is provided with functions of password protection and keyboard  
locking.  
The freezer must be unlocked before use: first press the “” key, the digits in the internal  
temperature display area will flash, and then press the “” or “” key to set the value as “06”, and  
finally press and hold the “Function selection” key for 8s to unlock the freezer.  
After installation, the freezer must be left still for over 24 hours before being energized.  
1. Initial state upon power-on:  
Turn on the battery switch at rear upper of compressor compartment, the SLT freezer is energized  
and enters into the “power-on” state, and all the parameters are the data before last power failure,  
and the display indicates actual temperature. The temperature setting is -126when the freezer  
leaves factory.  
If actual temperature is higher than temperature setting, delay for one minute, and then the  
compressor starts running. Meanwhile, if actual temperature is in the range of alarm temperature,  
light signals flash accompanied with audible buzz.  
2. Adjustment of set values  
2.1 Unlock  
2.2 “Function selection”: in unlock mode, press this key to select the desired function from “Internal  
temperature setting”, “High-temperature alarm setting” and “Low-temperature alarm setting”.  
“Internal temperature setting”: press the “Function selection” key continuously until the “Internal  
temperature setting” indicator light is on, at which time the display screen will indicate temperature  
setting and the digits flash, and then change temperature setting by pressing “” or “” to increase or  
decrease the value, specifically speaking, the value would increase or decrease by 1 if press “” or “”  
intermittently, while the value would increase or decrease rapidly and continuously by 1 if press and hold  
10-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
” or “”. After finishing the setting, if no key is pressed in 5s, the temperature value stops flashing,  
indicating that the value has been stored in the computer, otherwise the setting is invalid. The range of  
temperature setting is -126 ~ -150.  
In case of ambient temperature above 30(excluding 30), if the temperature setting is below -140,  
please adjust the internal temperature setting to -140; in case of ambient temperature below 28℃  
(including 28), please adjust the internal temperature setting to original set value.  
2.3 “High-temperature alarm setting”: press the “Function selection” key continuously until the  
“High-temperature alarm setting” indicator light is “on”, at which time the display screen will indicate  
temperature setting and the digits flash, and then change temperature setting by pressing “” or “” to  
increase or decrease the value, specifically speaking, the value will increase or decrease by 1 if press  
” or “” intermittently, while the value will increase or decrease rapidly and continuously by 1 if  
press and hold “” or “”. Once internal temperature is above high-temperature alarm setting, the  
alarm of SLT freezer sounds and the “High-temperature alarm” indicator light is on. The  
high-temperature alarm setting must not be above maximum temperature limit and 5lower than  
temperature setting.  
2.4 “Low-temperature alarm setting”: press the “Function selection” key continuously until the  
“Low-temperature alarm setting” indicator light is “on”, at which time the display screen will indicate  
temperature setting and the digits flash, and then change temperature setting by pressing “” or “” to  
increase or decrease the value, specifically speaking, the value will increase or decrease by 1 if press  
” or “” intermittently, while the value will increase or decrease rapidly and continuously by 1 if  
press and hold “” or “”. Once internal temperature is below low-temperature alarm setting, the  
alarm of SLT freezer sounds and the “Low-temperature alarm” indicator light is on. The low-temperature  
alarm setting must not be below minimum temperature limit and -5higher than temperature setting,  
otherwise the setting can not be performed.  
2.5 After finishing all the settings, if no operation is made in 5 minutes, the control board will be locked  
automatically, and these settings are valid.  
2.6 If any alarm condition occurs, all audible alarms will sound in such a way that it sounds for 15  
seconds and then pause for 1 minute until the condition is cleared.  
2.7 The setting of operating temperature should be as follows: first set the temperature at -126, and 24  
hours later, change to -135, and 7 hours later, change to required temperature, e.g. -140. When the  
freezer works normally in successive 3 days at temperature setting value, the 1st batch of samples may be  
loaded, but the volume of samples must not exceed 1/5 of the freezer’s effective capacity.  
11-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.8 If the freezer is required to reach its minimum temperature, e.g. –150, please do as follows: make  
sure that ambient temperature is below 25, and then set the temperature at –145and observe for 5~7  
days for stable working state and starting/stopping. If stable, change the temperature to -150.  
Note: Unless especially required (operating temperature -150), it is advisable to set the  
operating temperature above –145, otherwise that may have a negative influence on both efficiency  
and service life of the freezer.  
3. Condensate removal:  
In case of high ambient humidity, the anti-condensation switch at the rear of compressor compartment  
should be turned on to prevent condensate forming at the top opening of cabinet.  
4. If the SLT freezer is not to be used for a relatively long time, the battery switch at rear upper of  
compressor compartment should be turned off, otherwise the battery charge would be used up, and thus  
the battery could not be used any longer and recharged again.  
Tips:  
For any product may go wrong or fail to work, please follow the instructions below in order to avoid any loss  
resulting from faults of the freezer:  
1. The freezer must not be installed in a narrow space, and the door of room where it is installed must not be  
smaller than the freezer, but at least ensure normal handling to avoid such problems as difficult servicing, failure  
to repair it in time and damage of articles stored in the freezer, etc.  
2. The freezer should be attended by a qualified person and its operating status must be checked and recorded  
every day (check and record once every 2-4 hours). In case of any fault or power failure, the internal  
temperature would rise; if the fault may not be cleared in a short time, please take out stored articles and move  
them into other devices meeting the temperature requirement of these articles to prevent degeneration or  
damage.  
3. Before putting articles into the freezer, please make sure their temperature requirement is consistent with the  
freezer’s temperature range in order to prevent them from degeneration or damage due to the freezer’s failure to  
reach the temperature required by them.  
4. Due to refrigeration inertia, it is normal that there is certain difference between the temperature actually  
indicated and temperature setting (the lower temperature setting, the smaller temperature difference, whereas the  
larger temperature difference).  
12-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Defrosting · Non-use  
Defrosting  
1. After the SLT freezer is used for some time, a layer of frost may be formed; if too thick, it would  
influence the refrigeration effect and increase power consumption. Therefore, the freezer must be  
defrosted at certain interval, that is, upon the thickness reaching approximately 10mm.  
2. Before defrosting, take out all the articles stored in the freezer and move them into another freezer  
for preservation. For defrosting, open the door and place the freezer in a well-ventilated place.  
3. After defrosting, remove the resultant water with a soft cloth or sponge, wipe it with a dry cloth, and  
then turn it on, and finally load it with articles upon reaching the temperature setting.  
4. To speed up defrosting, a flat-bottom container filled with warm water below 50may be put inside  
of the cabinet.  
Non-use  
If the freezer is not to be used for a long time, power supply should be cut off and the freezer should  
be cleaned as described in previous sections. Open the door to have internal components dry, and then  
store it. The discarded freezer should be far away from fire sources and sent to dedicated place for  
disposal.  
Warning  
Do not defrost with an electric heater, a metal tool or an edged tool.  
13-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Cleaning  
Cleaning and maintenance  
The freezer should be cleaned after unpacking but before initial startup for commissioning, during  
normal use or before non-use.  
For safety, be sure to turn off the power switch before cleaning.  
The freezer should be cleaned with a soft cloth or sponge moistened with water or soap water  
(non-corrosive neutral detergents permitted), and then wiped with a dry cloth to prevent rust.  
During use, the dust on compressor compartment condenser, compressor and other pipelines should  
be removed with a soft cloth or a banister brush (soft hair brush) to keep optimal refrigeration effect.  
When cleaning, be careful not to damage pipelines and circuits.  
The condenser should be dusted out at least every 3 months.  
Do not clean the freezer with organic solvent, boiling water, detergent powder or acid, etc.  
Do not flush the freezer by directly spraying water onto it; do not use a scrub brush (hard hair brush)  
or steel wire brush for cleaning.  
If the freezer is not to be used for a long time, please turn it off, clean, dry and store it following the  
above instructions.  
14-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Q & A  
The following phenomena are not faults:  
In case of first use or startup, the freezer may work for a long time or give out loud noise, etc. That is  
not a fault but normal phenomenon.  
It is normal that the condenser fan starts and stops automatically according to ambient temperature.  
It is normal that there is fluid flow noise from the freezer arising from the cycle flow of refrigerant in  
its pipeline.  
15-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
After-sales Service  
Analysis on FAQ (frequently asked questions):  
If the condenser is dirty or clogged, it could not cool the compressed gas produced by the work of  
refrigerant compressor. When it is overheated, the compressor may be damaged. Therefore, the  
condenser must be cleaned on a regular basis to keep clean.  
If the freezer is installed close to a wall, it may lead to bad ventilation. In this case, once ambient  
temperature reaches 32, the condenser temperature would rise rapidly, seriously affecting the  
refrigeration system. Therefore, such measures must be taken as reinstallation, ventilation improvement  
or installation of air-conditioning system to ensure that ambient temperature is kept below 27℃  
(especially in hot summer or tropical zone).  
Unstable voltage or under-voltage may result in abnormal running of compressor. The rated voltage  
of compressor is 380V±10% (or 220V±10%), beyond which it is recommended to install a voltage  
stabilizer.  
If the freezer is installed in a closed room with air-conditioning system, it may be damaged due to a  
steep rise of ambient temperature in case of air-conditioner failure, especially on festivals and holidays  
or at night, so a monitoring system should be installed.  
Emergency measures  
In case of any fault or any other emergency,  
1. immediately move the samples stored in the freezer into a safe freezer in the vicinity.  
2. turn off power switch.  
3. contact and report the fault to local Haier After-Sales Service Center to have it clear soon.  
Common troubles and solutions  
The freezer does not work  
At first, check whether the air switch is tripped and whether power failure of external power  
network occurs, if the fault is not caused by the two reasons, check whether the AC contactor coil of  
compressor control board is energized; if energized, there is nothing wrong with the control board, and  
then check the circuit to compressor or compressor condition; if not, there is something wrong with the  
control board.  
16-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Solutions: If the problem lies in control board, replace it; if lies in compressor itself, replace it or its  
accessories.  
Unstable temperature  
Internal temperature does not reduce, but rises, e.g. rises to -80from -140.  
Cause 1: Dirty condenser or clogged air vent  
Solution: Clean the condenser  
Cause 2: Slowdown of fan speed  
Solution: Replace the fan  
Cause 3: The freezer is extremely close to a wall  
Solution: Move it far away from the wall for good ventilation  
Cause 4: The freezer is placed in a closed narrow space  
Solution: Install an air-conditioning system and improve ventilation condition  
Cause 5: The freezer is not left still for a sufficient time and is energized before its system reaches  
stable state, resulting in the problem of “oil beating” taken place in the compressor.  
Solution: Turn off the freezer, leave it still over 24 hours, and then turn it on, gradually reduce the  
temperature setting from -126by following corresponding operating instructions.  
Cause 6: The samples loaded at a time are over 1/5 of the freezer’s effective capacity, or the  
samples loaded are too hot (20higher than the temperature setting, e.g. the samples temperature  
is -106or above and the temperature setting is -126). In this case, the load of freezer greatly  
increases, so the compressor works for a prolonged time, and thus causes the temperature rise of  
compressor, steep reduction of refrigeration effect as well as imbalance between the cold energy  
provided by the refrigeration system and the loss of cold energy for the entry of ambient air, finally  
resulting in the rise of internal temperature.  
In this case, it is the fault of user for he/she uses the freezer as a quick freezer. However, the  
appliance is a SLT freezer for maintaining the low temperature of articles with only a little  
refrigeration and requires all the articles to be cooled down to basically equal temperature before  
being put into it. If the user fails to follow instructions described in this manual, the continuous  
17-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
running of compressor may be likely to cause compressor burn-out.  
Solution: Immediately have the compressor stop running; one hour later, i.e. after the compressor is  
moderately cooled down, adjust the temperature setting to a higher level to ensure that the  
compressor may stop running in 50 minutes; have the freezer operate at this temperature for 6~8  
hours, and then decrease the temperature by -10; have the freezer operate at this temperature for 8  
hours, and then decrease the temperature by -5~-10; in a word, decrease the temperature setting  
step by step until reaching the required temperature.  
Serious compressor noise  
Cause 1: It is normal that the noise level is higher than that of general home appliances and  
ultra-low temperature (ULT) freezers.  
Check: It is normal only if the system noise is not higher than 75dB.  
Solution: Do not use it in offices or laboratories where people stay for long.  
Cause 2: Fan motor  
Check: Check whether the noise is from compressor or fan motor.  
The noise from compressor is for the fan motor’s efficiency reduces and fails to cool the  
compressor effectively.  
Solution: Replace the fan motor  
Cause 3: The compressor pressure increases due to dirty or clogged condenser lowering  
refrigeration effect.  
Solution: Clean the condenser, and replace the compressor if necessary (serious damage).  
Cause 4: Over-high room temperature  
Over-high ambient temperature causes steep drop of refrigeration effect.  
Solution: Adjust room temperature and check noise level. Over-high noise level indicates serious  
compressor damage, in which case the compressor must be replaced.  
In case of any fault, be sure to immediately move the samples stored in the freezer into a proper device  
for safe preservation, and then turn off the freezer and contact Haier Headquarters or local After-sales  
Service Center to receive timely and fast service.  
18-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Telephone number of Haier Headquarters: 86-532-8893723688936023  
Warranty  
This series of products are under warranty in one year calculated from the date of purchase (subject  
to the invoice date; but calculated from the date on the “Installation and Commissioning Sheet” for those  
products purchased from dealers).  
As for the products out of warranty period, they will be charged as per Haier Uniform Service  
Charge Standard. For the replacement of parts, only cost and freight are charged and no charge on  
service is required to pay.  
The freezer enjoys Haier star-level service. If there is any problem, Haier will provide door-to-door  
service in 24 hours. Due to the particularity of both super low temperature (SLT) products and articles  
stored, the servicing and commissioning period is normally 15 days. For better use, we strongly  
recommend users to purchase a backup freezer or relevant backup systems in order to prevent influences  
on the normal storage of samples during servicing or in emergencies.  
The space below for pasting the invoice or Installation and Commissioning Sheet  
19-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Technical Parameter  
Model  
Overall  
(W×D×H)  
Internal  
(W×D×H)  
Compressor  
dimensions 1570×850×1060  
dimensions 667×462×650  
3.5HP  
DW-150W200  
Effective volume 200L  
Min.  
-150℃  
temperature  
Operating  
temperature  
range  
-126~-150℃  
Rated current  
2-phase 19A, 3-phase  
11A  
Rated voltage  
380V/50Hz  
220V/50Hz  
or Weight  
315kg  
Temperature decreasing curve reference  
DW-150W-200 curve [it is required to check the temperature decrease time (not including pre-cooling delay  
time) according to the ex-factory inspection requirements]  
20-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Packing List  
Qty.  
Description  
Warranty Card  
1
Operation Manual Frost scraper  
Plastic bag  
1
Model  
DW-150W200  
1
1
21-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Schematic Circuit Diagram  
22-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Schematic Refrigeration Diagram  
23-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Ambient Conditions  
The freezer is designed in consideration of following conditions and can be used in a safe way at least  
under the following conditions (on the basis of IEC 1010-1):  
1. Indoor use;  
2. Altitude not higher than 2000m;  
3. Ambient temperature within the range of 532;  
4. At a temperature not above 31, maximum relative humidity is 80%; the max. relative humidity  
decreases in a linear manner as temperature rises; max. relative humidity decreases to 50% upon  
reaching 40;  
5. The voltage fluctuation of main power supply is not over ±10% of its rated voltage;  
6. Additional fluctuation range of power supply voltage specified by the manufacturer;  
7. Up to the transient voltage of Equipment Installation Category (Over-voltage Category) II; for main  
power supply, the minimum and general category is Category II;  
8. Up to Pollution Class II specified in the Standard IEC664.  
24-24  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Graham Field Thermometer 1858 User Manual
Hamilton Beach Mixer ALL METAL DRINK MIXER User Manual
Harbor Freight Tools Window 91907 User Manual
Healthrider Treadmill HRAW56073 User Manual
Hoshizaki Freezer C 101BAH User Manual
Huffy Table Top Game 211011 User Manual
Husqvarna Lawn Mower 2246LS User Manual
Husqvarna Snow Blower 1130SBEXP User Manual
IBM Network Card PPC 1510PT User Manual
IBM Personal Computer 6268 User Manual