HP Hewlett Packard Network Router CFL 512 User Manual

Leycom® CFL-512 and Conduct NT  
CD Leycom September 2004  
The LEYCOM® Cardiac Function Lab CFL-512 facilitates the on-line  
measurement and analysis of ventricular pressure and volume signals in clinical  
setting. The pressure and volume signals are obtained using a combined  
conductance and pressure catheter. The CFL-512 acquires and displays these  
signals on-line and continuously. The signals are digitized, stored and can be  
subsequently played back and analyzed the users convenience.  
The Leycom® CFL-512 is a stand-alone device, which can be easily used in  
different clinical environments. The system, including the software for data  
acquisition and analysis, has been developed in close co-operation with a group  
of clinical users of Leycom® equipment.  
The Leycom® CFL-512 consists of a bedside 'patient module' and a console.  
The bedside 'patient module' is a state of the art, battery powered, data  
acquisition module. The module, to which the conductance catheter is  
connected, contains sophisticated circuitry, which ensures optimal patient safety  
and extended battery life.  
The 'patient module' is communicates with the Leycom® CFL-512 console via  
a fibre optic cable; this provides complete patient isolation from electrical  
hazard.  
The dedicated computer is Intel Pentium based, has a 40 Gb harddisk, enough  
for hundreds of hours of data acquisition, at 12 channels and 250 samples per  
second.  
With the Leycom® CFL-512 software, time-varying ventricular volume,  
pressure, ECG, a large number of hemodynamic variables and Pressure-volume  
loops can be displayed. The slope and intercept of the end-systolic pressure-  
volume relation (a measure of ventricular performance) are calculated, as well as  
other related variables.  
CD Leycom info 0310 V0440  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CFL-512  
Conduct NT – Analyse Recorded Data  
Conductance Catheter Settings  
Match the catheter with the actual heart size, exclude individual segments  
and optimize the dual field setting.  
Rho measurement  
The Rho cuvette is used to measure blood resistivity from a blood sample. Blood  
resistivity is required for the calibration of the volume measurements. The Rho  
cuvette is connected to the connector on the right side of the CFL-512. When  
the Rho cuvette is filled with 5 ml of blood the display will show the measured  
blood resistivity. By pressing the button next to the display this value is stored  
internally and used in the real-time volume calculation.  
Calibration  
Calibration of the volume signals from the conductance  
catheter is automatic. However, the 6 user defined signals  
connected to the external inputs, Pressure, ECG 2, Aux1,  
Aux2, Aux3 and Aux4 require initial calibration to enable the  
software to convert input signal voltages to user units (e.g.  
mmHg).  
Display  
Select display-mode and time-base for an optimal view of your chosen signals.  
View ‘Segmental Volumes’ to check and optimise the (segmental) volume  
signals. Use views with ‘PV-Loop’ to monitor the pressure-volume loops from  
beat to beat. Use views with ‘Indices’ to display real-time the calculated indices.  
Use ‘Dyssynchrony’ view to assess mechanical ventricular dyssynchrony.  
Acquire Data  
Start data acquisition by clicking ‘Acquire’. An audible signal every 1 s  
reminds the user that acquisition is in progress. Stop data acquisition by  
clicking acquire again. Enter a comment at the end of each acquisition run.  
CD Leycom info 0310 V0440  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CFL-512  
Conduct NT – Analyse Recorded Data  
Select data file  
Retrieve previously acquired data files for subsequent  
analysis.  
The comment and some file characteristics are displayed in  
the list.  
Pressure-volume analysis  
The end-systolic and end-diastolic PV points are indicated with markers (blue  
and green lines). Appropriate (user-selected) functions can be fitted through  
these points taken and varying pressures and volumes to calculate contractility  
and compliance.  
Print Report  
Create your own print reports with indices. Select the reports you want to print.  
Use the short-cut button  
to print the latest report.  
Indices  
Create your own indices-display by selecting the indices that are available in the  
list. Conduct NT has more than 50 parameters in his list of indices.  
Volume calibration  
Vc calculation (parallel conductance)  
Vc will be calculated automatically from the data acquired during a hypertonic  
saline injection.  
Alpha calculation  
To correct the underestimation of true volume, the results measured by the  
conductance catheter should be compared to the results of an independent  
cardiac output measurement, thermodilution, echo or angiography for example.  
Additional features in Conduct NT…..  
Interval measurement  
Several filters can be used  
Exporting data and indices to comma-separated file  
Beat to beat analysis  
Complete file management  
Signals can be individually scaled.  
CD Leycom info 0310 V0440  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CFL-512  
Highlights  
u
Hardware  
Software  
Console  
Conduct NT  
Real-time display  
Pentium, 1 GHz with 256 Mb RAM  
Dedicated hardware for:  
various display modes  
variable time-base  
Acquisition of external signals  
Communication with patient module  
Patient module battery management  
Monitor, NEC Multisync XP 1750  
Data storage, 40 GB harddisk  
250 Mb Zip drive  
Laser printer, HP LaserJet 2100  
Drip-proof keyboard with trackball and two buttons  
Several filters  
Data acquisition  
automatic file naming  
attached comments  
250 samples/sec, 12 bit accuracy  
8 internal (patient module) and 6 external channels  
Calibration of external signals  
conversion to user units  
optimize resolution  
Patient module  
Display of acquired data  
Calculation of calibration factors for conductance-  
volume  
Analysis of pressure-volume relations  
Calculation of hemodynamic indices  
Print/save graphs and own reports  
Printer settings  
All settings are software controlled  
Catheter connector: 16 pole Binder  
Battery powered to ensure patient safety  
More than 180 min continuous use when fully  
charged  
Communication with console via fiber optic cable  
AD conversion: 250 samples/sec, 12 bit  
Data export  
Patient safety  
Battery powered patient module  
Communication via fiber optic cable  
Inductive battery charger  
CD Leycom  
Argonstraat 116  
2718 SP Zoetermeer  
Tel. +31 79 3601780  
Fax. +31 79 3621743  
CD Leycom info 0310 V0440  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Grizzly Saw G1011Z User Manual
Haier Clothes Dryer HLP140E User Manual
Hamilton Beach Blender 54616C User Manual
Harbor Freight Tools Portable Generator 65309 User Manual
Heath Zenith Work Light SL 5503 User Manual
Honeywell Stud Sensor T7022A User Manual
Hotpoint Refrigerator FFP187B User Manual
Husqvarna Trimmer HVT40 User Manual
I Data Network Router ReRouter User Manual
IKEA Double Oven OV9 User Manual