Neb-u-Tyke®
®
Nebulizer Compressor
Item JB0112-070
User Manual
Read this manual before operating the nebulizer.
Save this manual for future reference.
Info: The most current version of this manual
Federal Law (USA) restricts this device to sale by or
on the order of a physician. This device should not
be used unless the operator has been instructed by
a qualified healthcare professional.
JB0112-070-INS-LAB-RevB14
Download from Www.Somanuals.com. All Manuals Search And Download.
INTRODUCTION
This user manual contains important information
and safety precautions for the John Bunn Neb-u-Tyke
Neb-a-Doodle Nebulizer Compressor.
Before using your Nebulizer Compressor, please read and
understand this entire user manual. Take special note of all
safety precautions marked DANGER and WARNING. Save
this user manual for future reference.
Intended Use
The Compressor Nebulizer System is designed to provide
a source of compressed air for aerosol therapy. It is used
in conjunction with a jet (pneumatic) nebulizer to produce
medicated aerosols for inhalation by pediatric and adult
patients with respiratory symptoms.
Indications for therapy include asthma, chronic bronchitis,
infection of the upper respiratory tract, chronic
obstructive pulmonary disease, and other respiratory
disorders in accordance with a physician’s prescription.
Do not use this product for any other purpose than the
preceding. This device can be used with adult or pediatric
patients under physician’s prescription.
The intended use of the Neb-a-Doodle Drawing Board is
drawing, writing, and doodling only.
JB0112-070-INS-LAB-RevB14
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Symbols Key
O
Off
I
~
On
Alternating current
IPX0
Enclosed equipment without protection against ingress of water
Attention (read the instructions)
II
Class II (not applicable for UL1431)
BF: This product is in accordance with the degree of protection against electric
shock for type BF equipment. (not applicable for UL1431)
Disposal of Electrical & Electronic Equipment (WEEE): This product should
not be treated as household waste, but submitted to an appropriate collection
point for the recycling of electrical and electronic equipment. For more detailed
information about the recycling of this product, please contact your local city
office, household waste disposal service, or the Graham-Field authorized
distributor from whom you purchased this product.
JB0112-070-INS-LAB-RevB14
4
Download from Www.Somanuals.com. All Manuals Search And Download.
IMPORTANT SAFETY PRECAUTIONS
When using electrical products, especially when children
are present, always follow basic safety precautions,
including the following:
DANGER: Indicates a potential hazard situation or unsafe
practice that, if not avoided, will result in death or serious
personal injury.
WARNING: Indicates a potential hazard situation or unsafe
practice that, if not avoided, could result in death or
serious personal injury.
CAUTION: Indicates a potential hazard or unsafe practice
that, if not avoided, could result in minor or moderate
personal injury.
s NOTICE: Indicates a potential hazard or unsafe practice
that, if not avoided, could result in product or property
damage.
Info: Provides application recommendations or other useful
information to ensure that you get the most from your
product.
READ ALL INSTRUCTIONS BEFORE USING
DANGER: To reduce the risk of electrocution:
1. Always unplug this product immediately after use.
2. Do not use this product while bathing, showering,
washing dishes, or close to water sources of any kind.
3. Do not place or store device where it can fall or be pulled
into a tub or sink.
4. Do not place device in or drop into water or other liquid.
5. Do not reach for a device that has fallen into water.
Unplug it immediately.
JB0112-070-INS-LAB-RevB14
5
Download from Www.Somanuals.com. All Manuals Search And Download.
WARNING: To reduce the risk of burns, electrocution, fire
or personal injury:
1. Never leave this device unattended when plugged in.
2. This product contains small parts that may present
a choking hazard to small children, and the tubing
may present a strangulation hazard. Always use close
supervision when this device is used by or near children
or those who require close supervision.
3. Children under the age of 3, or any patient who is unable
to use a mouthpiece properly under supervision, should
use a mask.
4. Use this device only as intended and described in this
manual. Do not use attachments or accessories not
recommended by GF Health Products, Inc.
5. Never operate this product if:
a. it has a damaged cord or plug,
b. it is not working properly,
c. it has been dropped or damaged, or
d. it has been dropped into water.
Return the product to your Graham-Field authorized
distributor for examination and repair.
6. Keep the cord away from heated surfaces.
7. Never block the air openings of this product or place it
on a soft surface, such as a bed or couch, where the air
openings may be blocked. Keep the air openings free of
lint, hair and the like.
8. Never drop or insert any object into any openings or hose.
9. Never use this device while drowsy or sleeping.
10. Do not use this device outdoors or operate where aerosol
(spray) products are being used.
11. Do not use this device in the presence of flammable
anesthesia or where oxygen or nitrous oxide is being
administered. This device has no AP or APG protection.
12. Connect this device to a properly grounded outlet only.
JB0112-070-INS-LAB-RevB14
6
Download from Www.Somanuals.com. All Manuals Search And Download.
13. This device has a polarized plug (one blade is wider
than the other). As a safety feature, this plug will fit into
a polarized outlet only one way. If the plug does not fit
fully into the outlet, reverse the plug. If it still does not fit,
contact a qualified electrician. Do not modify the plug in
any way. Do not attempt to defeat this safety feature.
14. Unplug the device before filling or cleaning the nebulizer.
15. This device is oil-free. Do not lubricate.
16. Do not remove the jet nebulizer while the device is operating.
17. Do not leave medication inside the jet nebulizer when it is
not in use.
18. Use the nebulizer intermittently; the maximum operating
period is 30 minutes. The nebulizer must stand unused for
30 minutes between uses.
19. The mouthpiece and masks are for use by one person
only; in order to avoid infection, do not share them.
20. Notice for California Customers- California Proposition 65
WARNING: This product contains a chemical known to the
State of California to cause cancer and reproductive or
developmental harm.
JB0112-070-INS-LAB-RevB14
7
Download from Www.Somanuals.com. All Manuals Search And Download.
PRODUCT DESCRIPTION
Features and Components
Please note nebulizer features and components, labeled in
the illustration below.
filter cap
air outlet connector
for tubing
ON/OFF switch
push button
tubing
reservoir tube
mouthpiece
nebulizer
housing
drawing
board
nebulizer cup
T adapter
adult mask
pediatric mask
shapes
slide tab
stylus
Neb-u-Tyke Neb-a-Doodle
Nebulizer Compressor
Info: Neb-a-Doodle replacement parts are available at
distributor.
JB0112-070-INS-LAB-RevB14
8
Download from Www.Somanuals.com. All Manuals Search And Download.
SETUP
Info: Before initial operation, clean the nebulizer
cup assembly following the instructions given in
MAINTENANCE/Cleaning the Nebulizer Cup.
WARNING: Ensure that the nebulizer compressor is
unplugged during setup.
1. Before each use, inspect the compressor and nebulizer
cup assembly for damage or wear; replace as needed.
2. Place the compressor on a table or other flat stable
surface (not on the floor), in an area that is clean and
dust-free. Be sure you can easily reach the controls
when seated.
3. Open the nebulizer medication cup: Twist the cap
counter-clockwise until it stops, then lift cap off.
4. Pour the prescribed medication into the nebulizer
medication cup. Do not fill with more than 5 ml per
treatment.
cap
5. Install cap on cup. Turn clockwise
cup
to tighten.
install cap
on cup
6. If using the mouthpiece for
treatment, proceed to step 7. If
using a mask, install the bottom
mask
part of the mask directly onto the
cap
top of the nebulizer medication cup
cap as shown at right. Proceed to
step 9.
install mask
on cap
T adapter
mouthpiece
7. Insert mouthpiece into large end of
T adapter.
assemble
mouthpiece
and T adapter
JB0112-070-INS-LAB-RevB14
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Info: The reservoir tube, shown in the illustration earlier in
this manual, acts as a holding tank for the mist, reducing the
amount lost during treatment. Its use is optional. If you wish
to use the reservoir tube, install it now on the small end of
the T adapter.
8. Install T adapter onto top of
T adapter
nebulizer medication cup cap.
cup
9. Connect the tubing to the adapter on
the bottom of the nebulizer medication
cup.
install
T adapter
on cap
10. Connect the tubing's other end
to the compressor air outlet
connector.
WARNING: Before use, check that all parts are securely
connected. Use of an improperly assembled nebulizer
could diminish or prevent adequate delivery of medication
and the effectiveness of the treatment.
JB0112-070-INS-LAB-RevB14
10
Download from Www.Somanuals.com. All Manuals Search And Download.
arrow
push button
OPERATION
Info: The drawing board must
be removed from the nebulizer
housing before use. To remove
the drawing board, push the
button and slide the drawing
1
2
board in the opposite direction from the cover arrow.
Nebulizer Operation
s NOTICE: The Neb-u-Tyke Neb-a-Doodle Nebulizer
Compressor is designed for intermittent use only. Do not
operate it continuously for more than 30 minutes for a
single use without turning it off and allowing a cooling
period of least 30 minutes.
1. Ensure that the ON/OFF switch is in the OFF position.
2. Plug the power cord into a properly grounded electrical
wall outlet.
3. Move the ON/OFF switch to the ON position to start the
compressor.
4. Place the nebulizer mouthpiece between teeth and
close mouth (or install the mask on face).
5. As aerosol starts flowing, inhale deeply and slowly
through mouthpiece (or mask).
WARNING: Never use a nebulizer with a clogged nozzle.
If the nozzle is clogged, the aerosol mist will be reduced,
altered or stopped, which will diminish or eliminate the
effectiveness of the treatment. If clogging occurs, replace
the nebulizer.
Info: If you need to interrupt treatment, move the ON/OFF
switch to the OFF position.
6. When the treatment is complete, move the ON/OFF
switch to the OFF position to turn off the compressor.
7. Unplug the compressor from power outlet.
8. Disconnect nebulizer from tubing.
9. Clean nebulizer as instructed in MAINTENANCE/
Cleaning the Nebulizer Cup.
JB0112-070-INS-LAB-RevB14
11
Download from Www.Somanuals.com. All Manuals Search And Download.
Drawing Board Operation
arrow
push button
1
2
To remove the drawing board, push the button and slide
the drawing board in the opposite direction from the cover
arrow.
To install the drawing board, insert the drawing board on
the side with the housing arrow that matches the cover
arrow, then push it all the way to secure it to the lock
button. This ensures the drawing board stays in place. To
prevent damage to the unit, ensure you do not insert it
from the other direction.
To draw on the drawing board, move the included stylus
pen, pointed end down, along the drawing board. To use
the included shapes, place them on the drawing board and
press lightly.
To erase the drawing board, move the slide tab over the
drawing board.
WARNING: This product contains small parts that may
present a choking hazard to small children. Always use
close supervision when this device is used by or near
children or those who require close supervision.
JB0112-070-INS-LAB-RevB14
12
Download from Www.Somanuals.com. All Manuals Search And Download.
Safety Overload
s NOTICE: Do not exceed 30 minutes of continuous
operation.
1. The motor of this device has a built-in thermal overload
protector. Should the motor overheat, the protector
will automatically shut off the motor. Should this occur,
move the ON/OFF switch to the OFF position and allow
the motor to cool down for approximately 30 minutes
before turning it on again.
2. If the overload protector shuts off the motor frequently,
you may have an unstable voltage situation.
3. If the device shuts down and cannot restart, it may
need to be replaced. Call your Graham-Field authorized
distributor immediately.
JB0112-070-INS-LAB-RevB14
13
Download from Www.Somanuals.com. All Manuals Search And Download.
MAINTENANCE
Cleaning the Compressor’s Outer Case
WARNING: Electrical shock hazard. Do not remove
the outer case of the device. All disassembly and
maintenance of the device must be performed by a
qualified service technician. Refer service to qualified
service personnel.
WARNING: This device does not require oil. Do not
attempt to lubricate any internal parts.
WARNING: Unplug device before cleaning. Do not
submerge in water for cleaning.
Wipe the compressor's outer case with a clean, damp cloth
weekly to keep it dust-free. Do not use any powdered type
cleaners or soap.
Cleaning the Nebulizer Cup
All parts of the disposable nebulizer, except tubing, should
be thoroughly cleaned after each use. If your physician
specifies a specific cleaning procedure, follow that
procedure.
WARNING: To prevent possible risk of infection from
contaminated medication, clean the nebulizer after each
treatment.
1. After each use, disassemble the nebulizer cup, T
adapter and mouthpiece completely, wash all parts in
warm soapy water, and rinse thoroughly. Allow to air
dry. Store in a clean, dry location.
2. Once daily, after washing as above, soak nebulizer
components in a fresh solution consisting of one part
warm water and one part white vinegar for 30 minutes
followed by a warm water rinse. Allow to air dry. Store
in a clean, dry location.
Cleaning the Drawing Board
Wipe the drawing board outside cover with a clean, damp
cloth weekly. Do not use any liquid, cleansers or soap on
any surface of the drawing board.
JB0112-070-INS-LAB-RevB14
14
Download from Www.Somanuals.com. All Manuals Search And Download.
Replacing the Filter
s NOTICE: Change the filter once every 30 days or when the
filter changes color. Additional filters can be purchased
authorized distributor.
1. Ensure that ON/OFF switch is in the OFF position. Open
the filter cap located next to the ON/OFF switch.
2. Remove the dirty filter with a small, pointed object.
Discard the filter.
3. Replace with a clean filter.
4. Replace the filter cap.
s NOTICE: Do not use cotton or other materials as a filter.
s NOTICE: Do not wash or clean the filter.
s NOTICE: Do not reuse the filter. Do not operate
compressor without a filter. Compressor damage can
result.
WARNING: The disposable nebulizer kit is for single
patient use only.
Storage
Keep the device and accessories dry. Avoid exposure to
direct sunlight.
JB0112-070-INS-LAB-RevB14
15
Download from Www.Somanuals.com. All Manuals Search And Download.
TROUBLESHOOTING
If your Neb-u-Tyke Neb-a-Doodle Nebulizer Compressor
fails to function, consult the following Troubleshooting
guide. If the problem persists, consult your Graham-Field
authorized distributor.
Problem
Device doesn’t
operate
Cause and Solution
Check if plug is properly fit into a grounded electrical outlet.
When device has been run continuously for more than 30 minutes
just before using, auto-shut-down may have been activated by the
built-in thermal protector; ensure that device is OFF for at least 30
minutes before next usage.
Nebulization is
weak
Check for proper electrical voltage.
Check tubing for blockage or air leakage at connection to Neb-u-Tyke
compressor or nebulizer cup; replace as needed.
Check that the nebulizer cup is properly assembled and not
damaged. If there is any damage, replace as needed.
Check if filter is dirty; replace filter as needed.
JB0112-070-INS-LAB-RevB14
16
Download from Www.Somanuals.com. All Manuals Search And Download.
SPECIFICATIONS
Specification
Description
Max pressure (compressor) 32.5 PSI
Max flow rate
Nebulization rate
Particle dimension
Capacity
Approximately 12 LPM
.33 ml/min
.5 - 5 microns
5 ml
Filter
Sponge
Dimensions (LxWxH)
Weight
10" × 6.5" × 4.2" (255 mm x 166 mm x 108 mm)
4 lb (1860 g)
Electrical requirement
Current
115VAC/60Hz
1.2A (rate)
Fuse
2.5A 250V
Power cord
Plug
Black, 7.5 ft (90") / 2,286 mm
2-prong type
56.5 dBA
Sound level
Environment
Temperature Operation: 41°F to 104°F (5°C to 40°C)
Storage: -40°F to 158°F (-40°C to 70°C)
Humidity
Operation: 30% to 95% RH
Storage: 10% to 95% RH
JB0112-070-INS-LAB-RevB14
17
Download from Www.Somanuals.com. All Manuals Search And Download.
LIMITED WARRANTY
SCOPE OF WARRANTY
GF Health Products, Inc. (“GF”) warrants to the original purchaser only that it
will replace or repair components, at GF’s sole discretion, that are defective
in material or workmanship under normal use and service. All warranties
are conditioned upon the proper use of the products strictly in accordance
with good commercial practice and applicable GF instructions and manuals,
including proper use and maintenance. To the extent that a component is
warranted by a third party, GF conveys all of its rights under that warranty to
the original purchaser, to the extent permitted.
This limited warranty shall only apply to defects that are reported to GF’s
customer service team within the applicable warranty period and which, upon
examination by GF or its authorized representative, prove to be a warranty
item. This limited warranty is not transferable.
The warranted components and time period are set forth below:
Compressor:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .five years
Accessories: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .six months
The applicable warranty period shall commence from date of shipment to
the original customer, unless there is an expiration date on the component in
which case the warranty shall expire on the earlier of warranty period or the
expiration date.
OBTAINING WARRANTY SERVICE
A GF Customer Service Representative must authorize warranty service.
Please contact the GF Customer Service department by calling 678-291-3207,
sending a fax request to 770-368-2386 or by e-mailing a request to cs@
grahamfield.com. Specific directions will be provided by the Customer Service
Representative. Failure to abide by the specific directions will result in denial
of the warranty claim.
EXCLUSIONS
The warranty does not cover and GF shall not be liable for the following:
1) Defects, damage, or other conditions caused, in whole or in part, by
misuse, abuse, negligence, alteration, accident, freight damage, tampering
or failure to seek and obtain repair or replacement in a timely manner;
2) Products which are not installed, used, or properly cleaned and maintained
as required in the official manual for the applicable product;
3) Products considered to be of a non-durable nature including, but not
limited to: casters, filters, fuses, gaskets, lubricants, and charts;
4) Accessories or parts not provided by GF;
5) Charges by anyone for adjustments, repairs, replacement parts, installation
or other work performed upon or in connection with such products which
are not expressly authorized in writing, in advance, by GF;
6) Any labor or shipping charges incurred in the replacement part installation
or repair;
7) Costs and expenses of regular maintenance and cleaning; and
8) Representations and warranties made by any person or entity other than
GF.
JB0112-070-INS-LAB-RevB14
18
Download from Www.Somanuals.com. All Manuals Search And Download.
ENTIRE WARRANTY, EXCLUSIVE REMEDY AND CONSEQUENTIAL
DAMAGES DISCLAIMER
THIS WARRANTY IS GF’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED. GF MAKES NO IMPLIED WARRANTIES OF
ANY KIND INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE.
IF ANY MODEL OR SAMPLE WAS SHOWN TO THE CUSTOMER, SUCH MODEL
OR SAMPLE WAS USED MERELY TO ILLUSTRATE THE GENERAL TYPE AND
QUALITY OF THE PRODUCT AND NOT TO REPRESENT THAT THE PRODUCT
WOULD NECESSARILY CONFORM TO THE MODEL OR SAMPLE IN ALL
RESPECTS.
THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF
THE DEFECTIVE PARTS. GF SHALL NOT BE LIABLE FOR AND HEREBY
DISCLAIMS ANY DIRECT, SPECIAL, INDIRECT, INCIDENTAL, EXEMPLARY OR
CONSEQUENTIAL DAMAGES, INCLUDING, BUT NOT LIMITED TO: DAMAGES
FOR LOSS OF PROFITS OR INCOME, LOSS OF USE, DOWNTIME, COVER,
OR EMPLOYEE OR INDEPENDENT CONTRACTOR WAGES, PAYMENTS AND
BENEFITS.
The warranties contained herein contain all the representations and
warranties with respect to the subject matter of this document, and supersede
all prior negotiations, agreements and understandings with respect thereto.
The recipient of this document hereby acknowledges and represents that it
has not relied on any representation, assertion, guarantee, warranty, collateral
contract or other assurance, except those set out in this document.
For additional information on this product or this warranty, please contact a
GF Customer Service Representative.
NOTES:
1) Additional terms and conditions may apply.
2) Freight claims must be notated on the Bill of Lading and must be made with
immediacy. The ICC regulations govern specific requirements for freight
claims. Failure to abide by those regulations may result in a denial of the
freight claim. GF will assist you in filing the freight claim.
3) Claims for any short shipment must be made within thirty (30) days of the
invoice date.
JB0112-070-INS-LAB-RevB14
19
Download from Www.Somanuals.com. All Manuals Search And Download.
Manufactured for:
GF Health Products, Inc.
2935 Northeast Parkway
Atlanta, Georgia 30360 USA
Telephone: 770-368-4700
Fax: 770-368-2386
Made in China
Download from Www.Somanuals.com. All Manuals Search And Download.
Neb-u-Tyke®
®
Compresor de Nebulizador
Modelo JB0112-070
Manual de Usuario
Lea antes de operar el nebulizador. Guarde este
manual para referencia en el futuro.
Info: La más reciente versión de este manual podrá
La Ley Federal (USA) restringe este aparato para
ventas por la orden de su médico. Este aparato no
debe ser utilizado si el operador no ha sido adiestrado
por profesional de cuido de salud cualificado.
JB0112-070-INS-LAB-RevB14
Download from Www.Somanuals.com. All Manuals Search And Download.
Este producto ha sido examinado y aprobado con los siguientes
estándares:
UL1431
EN 60601-1
CONTENIDOS
INTRODUCCIÓN...................................................................................................3
USO DEL COMPRESOR DE NEBULIZADOR..............................................3
SÍMBOLOS ....................................................................................................4
DIRECTIVAS IMPORTANTES DE SEGURIDAD .................................................5
DESCRIPCIÓN DEL PRODUCTO........................................................................8
CARACTERÍSTICAS Y COMPONENTES ....................................................8
ENSAMBLAJE......................................................................................................9
OPERACIÓN....................................................................................................... 11
OPERACIÓN DEL COMPRESOR DE NEBULIZADOR.............................. 11
OPERACIÓN DE LA TABLA DE DIBUJAR................................................12
SEGURIDAD EN CASO DE SOBRECARGA .............................................13
MANTENIMIENTO .............................................................................................14
LIMPIEZA DE LA CARCASA DEL COMPRESOR .....................................14
LIMPIEZA DE LA TAZA DEL NEBULIZADOR ...........................................14
LIMPIEZA DE LA TABLA DE DIBUJAR .....................................................14
INSTRUCCIONES PARA EL CAMBIO DEL FILTRO DE AIRE .................. 15
ALMACENAJE ............................................................................................15
DETECTANDO PROBLEMAS............................................................................16
ESPECIFICACIONES .........................................................................................17
GARANTÍA LIMITADA........................................................................................18
GF Health Products, Inc. no es responsable de errores tipográficos. Para
obtener la información más actualizada y vigente relativa a los envases,
garantías, productos y especificaciones, incluyendo la versión más
actualizada de estas instrucciones, por favor visite nuestro sitio web en
Graham-Field, John Bunn, Neb-u-Tyke y Neb-a-Doodle son marcas registradas
de GF Health Products, Inc. © 2009 GF Health Products, Inc.
JB0112-070-INS-LAB-RevB14
2
Download from Www.Somanuals.com. All Manuals Search And Download.
INTRODUCCIÓN
Este manual del usuario contiene información importante
y precauciones de seguridad para el John Bunn Neb-u-Tyke
Neb-a-Doodle Compresor de Nebulizador. Antes de usar
su Compresor de Nebulizador, favor de leer y entender
el manual de usuario. Tome nota especial en todas las
precauciones marcadas PELIGRO y ADVERTENCIA. Guarde
este manual de usuario para referencia en el futuro.
Uso del Compresor de Nebulizador
El Neb-u-Tyke Neb-a-Doodle Compresor de Nebulizador
es diseñado para proveer una fuente de aire comprimido
para la terapia de aerosol. Se usa en conjunto con
un nebulizador de chorro (neumático) para producir
aerosoles de medicamento para inhalación por pacientes
adultos y pediátricos con síntomas respiratorios.
Indicaciones para terapia incluye asma, bronquitis crónica,
infección de la zona superior respiratorio, enfermedad
pulmonar obstructiva crónica y otros desordenes que va
de acuerdo con la receta de su médico. No debe usar este
producto por cualquier otro propósito, que lo descrito
arriba. Este aparato puede ser usado con pacientes adultos
y pediátricos con la receta de su médico.
El uso entendido del Neb-a-Doodle Tabla de Dibujo es
dibujar, escribir y garabatear solamente.
JB0112-070-INS-LAB-RevB14
3
Download from Www.Somanuals.com. All Manuals Search And Download.
Símbolos
O
Apagado
I
~
Encendido
Corriente alternando
IPX0
El equipo no tiene protección contra sumergiendo en agua
Atención (lea las instrucciones)
II
Clase II (no aplicable para UL1431)
BF: Este producto está en concordancia con el grado de protección sobre el
riesgo de descarga eléctrica por equipo de tipo BF (no aplicable por UL1431).
Disposición de Equipo Eléctrico y Electrónico (WEEE): Este producto no debe ser tratado
como residuos domésticos, pero sometidos a un sitio de colección para reciclaje de equipo
eléctrico y electrónico. Para más información detallada sobre el reciclaje de este producto,
llame a su oficina de cuidad local, un servicio de disposición de residuos domésticos o a su
distribuidor autorizado de Graham-Field de donde compró este producto.
JB0112-070-INS-LAB-RevB14
4
Download from Www.Somanuals.com. All Manuals Search And Download.
DIRECTIVAS IMPORTANTES DE SEGURIDAD
PELIGRO: Indica una situación de peligro o una practica
insegura que, si no es evitada, resultará en muerte o
heridas serias a la persona.
ADVERTENCIA: Indica una situación de peligro o una
practica insegura que, si no es evitada, puede resultar en
muerte o heridas serias a la persona.
PRECAUCIÓN: Indica una situación de peligro o una
practica insegura que, si no es evitada, podría dar lugar a
daños corporales de menor importancia.
s AVISO: Indica una situación de peligro o una practica
insegura que, si no es evitada, podría dar lugar a daños
de menor importancia al producto o daños materiales.
Info: Proporciona recomendaciones para la aplicación y
otra información útil para asegurar que obtengas lo más que
pueda del producto.
LEE TODAS LAS INSTRUCCIONES ANTES DE UTILIZAR
PELIGRO: para reducir el riesgo de electrocución.
1. Siempre debe desconectar este producto inmediatamente
después de usar.
2. No debe usar durante el baño, lavando platos, y cerca de
cualquier fuente de agua.
3. No posicionar o almacenar donde pueda caer a la bañera
o al lava manos.
4. No debe introducir en agua u otro liquido. No debe
recoger producto sumergido en agua. Desconecte
inmediatamente.
JB0112-070-INS-LAB-RevB14
5
Download from Www.Somanuals.com. All Manuals Search And Download.
ADVERTENCIA: Para reducir el riesgo de quemadura,
electrocución, fuego o heridas a la persona.
1. Este producto nunca debería ser desatendido mientras
esté conectado.
2. Este producto contiene piezas pequeñas que presenta
riesgos de asfixia para niños, y la manga puede presentar
riesgo de estrangulación. Supervisión cercana es
necesaria cuando este producto esté utilizado cerca de
niños o personas con impedimentos.
3. Niños menores de 3 años, y cualquier paciente quien no
puede usar la boquilla correctamente bajo supervisión,
debe usar la máscara.
4. Utilice este producto solamente como indicado en el
manual de usuario. No utilice aditamentos o accesorios
no recomendados por GF Health Products, Inc.
5. Nunca utilice este producto si
a. el enchufe o el cable está dañado,
b. no funciona adecuadamente,
c. ha sufrido caída o daño, o
d. ha sido mojado o sumergido en agua.
Devuelve este producto al centro de servicio para examen
y/o arreglos.
6. Mantenga el cable lejos de superficies calientes.
7. Nunca obstruya entradas de aire del producto o utilice
en una superficie suave tal como cama o sillón donde las
entradas del aire pueden ser bloqueadas. Mantenga las
entradas libres de pelusa, cabello, o materiales de este
tipo.
8. Nunca inserte objetos en cualquier entrada o manga.
9. Nunca utilice soñoliento o dormido.
10. No utilice al aire libre, ni opere donde se utilicen
productos de aerosol.
11. No utilice en presencia de anestesia, oxígeno, oxido
nítrico, o cualquier gas inflamable mientras estos estén
siendo administrado. Este aparato no tiene protección AP
o APG.
JB0112-070-INS-LAB-RevB14
6
Download from Www.Somanuals.com. All Manuals Search And Download.
12. Conecte este producto solamente en un enchufe de pared
correcto.
13. Este aparato tiene un enchufe polarizado (un aspa es más
ancha que la otra). Para reducir el riesgo de electrocución,
este enchufe es diseñado para poder ser posicionado
de una sola manera en una toma eléctrica polarizada. Si
el enchufe no entre completamente en la toma, invierta
el enchufe. Si no funciona de este modo, contacte un
electricista cualificado. No modifique el enchufe de ninguna
manera.
14. Desconecte el aparato antes de llenar el nebulizador.
15. Este aparato no utiliza aceite. No lubrique.
16. No debe remover el nebulizador de chorro cuando el
aparato está operando.
17. No debe dejar medicamento en el nebulizador cuando no
esté en uso.
18. Use el nebulizador por periodos: el periodo de operación
máximo es 30 minutos. El nebulizador no debe ser usado
por 30 minutos entre usos.
19. La boquilla y las máscaras son para uso de una sola
persona; para evitar infección no debe compartirlos.
20. Aviso para los clientes de California - Proposición 65 de
California ADVERTENCIA: Este producto contiene una
sustancia química conocida en el estado de California por
causar cáncer y daños reproductivos o del desarrollo.
JB0112-070-INS-LAB-RevB14
7
Download from Www.Somanuals.com. All Manuals Search And Download.
DESCRIPCIÓN DEL PRODUCTO
Características y componentes
Por favor de tomar nota de las características y los
componentes del nebulizador descritos en la ilustración
abajo:
tapa del filtro
conector del
enchufe de aire
interruptor de ON/OFF
botón
manga
manga de reserva
boquilla
compartimento
de nebulizador
tabla de
dibujar
taza del nebulizador
adaptador T máscara adulta
máscara pediátrica
formas
lengüeta
lápiz
Neb-u-Tyke Neb-a-Doodle
Compresor de Nebulizador
Info: Neb-a-Doodle piezas reemplazadas están disponibles
Graham-Field.
JB0112-070-INS-LAB-RevB14
8
Download from Www.Somanuals.com. All Manuals Search And Download.
ENSAMBLAJE
Aviso: Antes de operación inicial, limpie la taza del
nebulizador siguiendo las instrucciones escritos en la sección
Mantenimiento/Limpieza de la Taza del Nebulizador.
ADVERTENCIA: Antes de conectar el cordón de
electricidad, asegure que el interruptor de ON/OFF está en
la posición OFF.
El enchufe también sirve para desconectar el aparato.
1. Antes de uso, examine el compresor y taza de
nebulizador para daño; reemplace si es necesario.
2. Posicione el aparato en una superficie horizontal para
que los controles estén con fácil acceso. No debe usar
este aparato en el piso.
3. Remueva la tapa de la taza del nebulizador.
4. Inserte medicamento prescrito por su médico.
tapa
5. Reúne taza y tapa.
taza
ensamblaje de
taza y tapa
6. Si usa la boquilla para el tratamiento,
proceda al paso 7. Si usa la máscara,
máscara
instale la parte inferior de la máscara
directamente sobre la tapa de la taza
tapa
del nebulizador como demostrado a
la derecha, entonces proceda al paso
9.
instale la máscara
en la tapa
adaptador T
7. Conecte la boquilla a la parte
boquilla
grande del adaptador T.
ensamblaje de
adaptador T
y boquilla
JB0112-070-INS-LAB-RevB14
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Info: La manga de reserva, en la ilustración superior del
manual, es como un tanque de retención para reducir la
cantidad de medicamento perdido durante tratamiento. Es
opcional. Si desea usar la manga de reserva, posicione en la
parte inferior del adaptador T.
8. Instale el adaptador T sobre la
adaptador T
tapa de la taza del nebulizador.
taza
9. Conecte la manga al adaptador
a la parte inferior de la taza
nebulizador.
instale adaptador T
en tapa de la taza
10. Conecte la otra parte de la manga al conector de mango
del compresor.
ADVERTENCIA: Antes de uso, verifique que todas las
piezas están conectados correctamente. Uso de un
nebulizador armado incorrectamente puede disminuir
o impedir reparto adecuado de medicamento y la
efectividad del tratamiento.
JB0112-070-INS-LAB-RevB14
10
Download from Www.Somanuals.com. All Manuals Search And Download.
flecha
botón
OPERACIÓN
Info: La tabla de dibujar tiene
que ser apartado del nebulizador
antes de uso. Para quitar la tabla
de dibujar, empuje el botón y
1
2
deslice la tabla de dibujar en la dirección opuesto de la flecha
en la tapa.
Operación del Compresor de Nebulizador
s AVISO: El Compresor de Nebulizador está diseñado
para uso solamente por periodos. No debe operarlo
continuamente por más de 30 minutos en un uso sin
apagarlo y dejar que se enfríe por lo menos 30 minutos.
1. Asegurase que el interruptor eléctrico esté en la
posición “OFF”.
2. Conecte el cordón eléctrico en la toma eléctrica apropiada.
3. Encienda el compresor posicionando el interruptor a ON.
4. Acomode la boquilla del nebulizador entre los dientes y
cierre la boca (o instale la máscara en la cara).
5. Cuando el aerosol empiece a fluir, necesita inhalar
profundamente y despacio a través de la boquilla o la
máscara.
ADVERTENCIA: Nunca debe usar el nebulizador con un
inyector tapado. Si el inyector está tapado, la nube de
aerosol será reducido, cambiado o parado, que puede
diminuir o eliminar la efectividad del tratamiento. Si se
tapa, reemplace el nebulizador.
Info: Si el tratamiento necesita ser interruptado, posicione el
interruptor a la posición OFF.
6. Cuando el tratamiento haya terminado, apague el
compresor posicionando el interruptor en la posición OFF.
7. Desenchufe el aparato de la toma eléctrica.
8. Desconecte el nebulizador del mango.
9. Limpie el nebulizador siguiendo las instrucciones
escritos en la sección Mantenimiento/Limpieza de la
Taza del Nebulizador.
JB0112-070-INS-LAB-RevB14
11
Download from Www.Somanuals.com. All Manuals Search And Download.
Operación de la Tabla de Dibujar
flecha
botón
1
2
Para quitar la tabla de dibujar, empuje el botón y deslice la
tabla de dibujar en la dirección opuesto que la flecha en la
tapa.
Para instalar la tabla de dibujar, inserte la tabla de dibujar
en el lado con la flecha en el compartimento que machea
la flecha en la tapa, y empuje hasta el final para asegurarlo
con el botón de cerrar. Esto asegure que la tabla de dibujar
se queda en su sitio. Para prevenir daño al aparato,
asegure que no inserta la tapa por la otra dirección.
Para dibujar en la tabla de dibujar, mueve el lápiz, punto
hacia abajo, sobre la tabla de dibujar. Para usar las
formas incluidos, coloca sobre la tabla de dibujar y apreté
suavemente.
Para borrar la tabla de dibujar, mueve la lengüeta sobre la
tabla de dibujar.
ADVERTENCIA: Este producto contiene partes pequeñas
que pueden presentar un peligro de asfixiar para niños.
Siempre utilice supervisión cerca cuando usando
este aparato cerca de niños o los quienes requieren
supervisión cerca.
JB0112-070-INS-LAB-RevB14
12
Download from Www.Somanuals.com. All Manuals Search And Download.
Seguridad en caso de sobrecarga
s AVISO: No debe exceder más de 30 minutos de operación
continuo.
1. Este compresor contiene un dispositivo de protección
termal. Si el motor comienza a sobrecalentarse, se
apagará automáticamente. Si esto ocurre, posicione el
interruptor a la posición OFF y permita que el motor
se enfrié por 30 minutos antes de recomenzar el
compresor.
2. Si el protector apaga el motor frecuentemente, podría
tener una situación de voltaje inestable.
3. Si el aparato se apaga y no se enciende, puede ser
necesario reemplazarlo. Llame a su proveedor
inmediatamente.
JB0112-070-INS-LAB-RevB14
13
Download from Www.Somanuals.com. All Manuals Search And Download.
MANTENIMIENTO
Limpieza de la Carcasa del Compresor
ADVERTENCIA: Riesgo de electrocución. No remueve
carcasa externa de este aparato. Todo desensamblaje y
mantenimiento de este aparato debe ser realizado por un
técnico de servicio cualificado.
ADVERTENCIA: Este aparato no necesita aceite. No
lubrique piezas internas.
ADVERTENCIA: Desenchufe el aparato antes de limpiar.
No sumerja bajo agua.
Frote con un paño húmedo cada semana. No utilice
detergentes en polvo o abrasivos que pueden dañar la
carcasa.
Limpieza de la Taza del Nebulizador
Todas las piezas del nebulizador desechable, excepto la
manga, deben ser limpiados después de cada uso. Si su
médico le da un procedimiento especifica de limpieza, siga
ese procedimiento.
ADVERTENCIA: Para prevenir riesgo de infección
causado por medicamento contaminado, la limpieza del
nebulizador se recomienda después de cada tratamiento.
1. Después de cado uso, desensamble la taza
del nebulizador, el adaptador T y la boquilla
completamente. Lave todas las piezas en agua tibia con
jabón y enjuague bien.
2. A diario, después del paso anterior, remoje los
componentes del nebulizador en una solución
consistiendo de una parte agua y una parte vinagre por
30 minutos, seguido por un enjuague con agua tibia.
Limpieza de la Tabla de Dibujar
Limpie la tabla de dibujar con una toalla limpia y con
humedad seminal. No debe utilizar líquidos, productos de
limpieza o jabones in cualquier superficie de la tabla.
JB0112-070-INS-LAB-RevB14
14
Download from Www.Somanuals.com. All Manuals Search And Download.
Instrucciones para el cambio del filtro de aire
Aviso: Cambie el filtro cada 30 días o cuando el filtro
cambia color. Filtros adicionales pueden ser comprados en
Graham-Field.
1. Remueva la tapa del filtro de aire localizado al lado del
interruptor ON/OFF.
2. Remueva el filtro sucio con un objeto punzante.
3. Reemplace el filtro nuevo.
4. Reemplace la tapa del filtro.
s AVISO: No utilice algodón o cualquier material como filtro.
s AVISO: No lave o limpie el filtro.
s AVISO: No reutilice el filtro. No opere sin filtro. El
compresor puede ser dañado por este uso.
ADVERTENCIA: El kit del nebulizador desechable es para
uso solamente de una persona.
Almacenaje
Mantenga el aparato y los accesorios secos. Evite sol
directamente.
JB0112-070-INS-LAB-RevB14
15
Download from Www.Somanuals.com. All Manuals Search And Download.
DETECTANDO PROBLEMAS
Si su Neb-u-Tyke Neb-a-Doodle Compresor de Nebulizador
falla, consulte a la guía de Detectando Problemas debajo.
Si el problema continua, llame a su distribuidor autorizado
de Graham-Field.
Problema
El aparato no
funciona
Causa y Solución
Verifique si el enchufe está conectado correctamente en la toma
eléctrica apropiada.
Cuando el aparato ha estado corriendo continuamente por más de 30
minutos, el protector termal interno puede ser activado para que el
aparato se enfríe por 30 minutos, antes del próximo uso.
Nebulización es Verifique que tiene el voltaje eléctrico apropiado.
débil
Verifique la manga para bloqueos o escapes de aire en la conexión al
compresor o la taza de nebulizador, reemplace si es necesario.
Verifique que la taza del nebulizador está armado bien y no dañado. Si
hay daño, reemplace si es necesario.
Verifique si el filtro está sucio, reemplace el filtro.
JB0112-070-INS-LAB-RevB14
16
Download from Www.Somanuals.com. All Manuals Search And Download.
ESPECIFICACIONES
Especificación
Presión máxima del
compresor
Descripción
32.5 PSI
Flujo máximo
Tasa de nebulización
Tamaño de partícula
Capacidad
Aproximadamente 12 LPM
0.33 ml/min
.5 - 5 microns
5 ml
Filtro
Esponja
Dimensión
(largo x anchura x altura)
10" × 6.5" × 4.2"(255 mm x 166 mm x 108 mm)
Peso
4 lb (1860 g)
Requísito eléctrico
Corriente
115VAC/60Hz
1.2A (tasa)
Capacidad de fusible
Cordón eléctrico
Enchufe
2.5A 250V
Blanco, 7.5 ft (90") / 2,286 mm
De tipo de 2 Aspa
56.5 dBA
Nivel de sonido
Ambiente
Temperatura Operación: 41°F a 104°F (5°C to 40°C)
Almacenaje: -40°F to 158°F (-40°C to 70°C)
Humedad
Operación: 30% a 95% RH
Almacenaje: 10% a 95% RH
JB0112-070-INS-LAB-RevB14
17
Download from Www.Somanuals.com. All Manuals Search And Download.
GARANTÍA LIMITADA
AMBITO DE GARANTÍA
GF Health Products, Inc. (“GF”) garantía al comprador original o arrendatario
(cual quiera es el “Cliente”) solamente, que va reemplazar o reparar
componentes, a la decisión solo de GF, cuales son defectos en material o
calidad utilizado normalmente y con servicio. La garantía es para partes
solamente y no incluye fuerza laboral o el costo de transporte. Esta garantía
limitada no es transferible. Todas garantías son condicionales por el uso
apropiado del producto estrictamente en acuerdo con las instrucciones
aplicables de GF, incluyendo uso apropiado y mantenimiento. Producto
próvido a usuarios como réntales deben ser serviciados por el proveedor
y inspeccionado antes de entrega. Cada usuario debe ser preparado en la
operación y seguridad de este producto antes de uso. Dentro las reglas
normales en este documento, el John Bunn JB0112-070 Compresor Nebulizador
y sus componentes a estar libre de defectos de materiales o mano de obra
según listado debajo:
Compresor: cinco años
Accesorios: seis meses.
El período de garantía se comienza en la fecha de transporte al Cliente a
menos que haya una fecha de expiración en el componente en cual la garantía
se expira en la período de garantía o la fecha de expiración.
OBTENIENDO SERVICIO DE GARANTÍA
La garantía limitada solamente aplica a defectos reportados al Distribuidor
de quien el Cliente compro el producto y quienes ellos en turno notifica al
equipo de Servicio de Cliente de GF dentro el período de garantía aplicable en
cuanto será examinado por GF o un representante autorizado para probar el
producto sobre garantía. Falta de seguir las instrucciones específicas resultara
en rechazo de la reclamación de garantía. Si no hay Distribuidor, debes llamar
a GF directamente llamando 678-291-3207, mandando fax a 770-368-2386 o
por email a cs@grahamfield.com. Direcciones específicos se van a dar por el
Representante de Servicio De Clientes.
EXCLUSIONES
La garantía no cubre ni GF será responsable de lo siguiente:
1) Defectos, daño u otras condiciones causadas por mal uso, abuso,
negligencia, alteración, accidente, daño de transporte o falta de buscar y
obtener reparo o repuesto en manera puntual;
2) Productos con los números seriales quitados o desfigurados;
3) Productos no instalados, utilizados or mantenidos correctamente como
requerido en los manuales oficiales del Producto; y
4) Cargos de Labor o transporte incurridos en el regreso o instalación de
partes de reemplace;
5) Productos considerados ser consumibles como dicho en este documento;
6) Asesorías o partes no próvidos por GF;
7) Cargas por otras personas por ajustamientos, reparaciones, partes de
reemplace, instalación u otro trabajo hecho en conexión con este Producto
no autorizado en escrito en avanza por GF;
8) Cargas y gastos de mantenimiento y limpieza regular; y
9) Representaciones y garantías hechos por otras personas o entidades no
GF.
JB0112-070-INS-LAB-RevB14
18
Download from Www.Somanuals.com. All Manuals Search And Download.
GARANTÍA ENTERO, REMEDIO EXCLUSIVO Y DESCARGO DE
DAÑOS CONSIGUIENTE
ESTA GARANTIA ES LA UNICA GARANTIA DE GF Y ES EN VEZ DE TODAS
OTRAS GARANTIAS, EXPRESIDAS O IMPLICITADAS. GF NO HACE GARANTIAS
ENTENTIDOS DE CUAL QUIER MANERA INCLUYENDO GARANTIAS
ENTENDIDOS DE COMERCIALIZACION O IDONEIDAD PARA UN PROPOSITO
PARTICULAR. SI UN MODELO Y MUESTRA SE LE ENSENO A UN CLIENTE,
TAL MODELO O MUESTRA FUE UTILIZADO SOLAMENTE PARA ILUSTRAR
EL TIPO GENERAL Y CALIDAD DEL PRODUCTO Y NO PARA REPRESENTAR
QUE EL PRODUCTO SE CONFORME CON ESE MODELO Y MUESTRA EN
TODOS RESPETOS. TODA RESPONSABILIDAD DE GF POR CUAL QUIER
PRODUCTO Y SERVICIO PROVIDO ES LIMITADA AL COSTO DEL PRODUCTO
EN LA RECLAMACION. EN NO EVENTO EN CONTRACTO, INDEMNIZACION,
GARANTIA, NEGLIGENCIA, RESPONSABILIDAD ESTRICTA O OTRO QUE GF
SERA RESPONSABLE POR CUAL QUIER DANO DIRECTO, ESPECIAL, INDIRECTO,
INCIDENTAL, EJEMPLAR O CONSIGUIENTES INCLUYENDO PERO NO LIMITADO
A: DANOS FOR PERDIDA DE BENEFICIO O GANANCIA, PERDIDA DE USO,
INACTIVO, O SALARIOS DE EMPLEADOS O TRABAJADORES CONTRACTORES,
PAGOS O BENEFICIOS.
Esta garantía te da derechos legales específicos. Podrás tener derechos
adicionales que varían estado por estado.
JB0112-070-INS-LAB-RevB14
19
Download from Www.Somanuals.com. All Manuals Search And Download.
Fabricado para:
GF Health Products, Inc.
2935 Northeast Parkway
Atlanta, Georgia 30360 USA
Telephone: 770-368-4700
Fax: 770-368-2386
Fabricado en China
Download from Www.Somanuals.com. All Manuals Search And Download.
|