LF1031 HYDRAULIC PATIENT LIFT
LF1041 BATTERY POWERED PATIENT LIFT
USER MANUAL
LF1031-INS-LAB-RevD12
© 2012 GF Health Products, Inc.
Read this manual before operating your Patient Lift.
Save this manual for future use.
The most current version of this manual can be found online at
Download from Www.Somanuals.com. All Manuals Search And Download.
1 INTRODUCTION
Congratulations on your purchase of a Lumex patient lift. The following pages will provide you
with important safety and operating instructions on the use of patient lifts, slings, and accessories,
as well as maintenance and warranty information. Read this manual carefully before operating
your patient lift and refer to it as often as needed. Consult your Lumex dealer and/or healthcare
professional with any questions or concerns regarding safe and effective techniques for operating
your patient lift.
The Lumex Patient Lift is intended to aid in the transfer of an individual who is unable to
independently transfer between bed, chair, wheelchair, and/or commode.
DO NOT attempt any transfer without prior approval of the patient’s healthcare provider. With
their assistance, select the sling that best serves the needs, abilities, comfort and safety of the
person being lifted. It is important that you use correct slings adapted to the body shape and
weight and ability of the patient and the specific type of transfer needed.
Upon receipt of the patient lift, and prior to each use, inspect it closely to ensure that nothing is
damaged, there are no loose or broken parts, that all patient lift parts are in the proper place, etc.
Immediately replace any worn or broken components. Your patient lift should receive maintenance
on a regular schedule and should be inspected daily for proper operation. Contact your Lumex
dealer for service and repairs.
Before lifting anyone, thoroughly read the instructions in this manual. The operator should
practice the entire lifting procedure several times with proper supervision and with a capable
individual acting as patient. It is extremely important to know how a patient will feel in the patient
lift. We suggest that the operator try all the various lifting slings and equipment until the operator
is confident with how the patient lift operates. The lifting procedures should be explained to the
patient before the actual lifting. Remember that safety requires the constant attention of the lift
occupant and the operator. Use extreme care and common sense when operating a patient lift.
This guide covers the use of the LF1031 Hydraulic Patient Lift and the LF1041 Battery Powered
Patient Lift. Keep in mind that the basic safety procedures included in this manual are to be used
as a guide only. You may find it necessary to develop your own unique methods for safely utilizing
the patient lift. Again, consult your healthcare professionals for their recommendations and never
hesitate to ask for their assistance.
Info: Before attempting to assemble the patient lift, refer to ASSEMBLY INSTRUCTIONS on
page 8. Contact your Lumex dealer if additional assistance is needed.
Info: In this manual, the person being lifted is referred to as the “patient” or “occupant”
and the person operating the patient lift is called the “operator” or “attendant”.
Please note the following special statements, used throughout this manual, and their significance:
WARNING: Indicates a potential hazard situation or unsafe practice that, if not avoided,
could result in death or serious personal injury.
CAUTION: Indicates a potential hazard situation or unsafe practice that, if not avoided,
could result in minor or moderate personal injury.
s NOTICE: Indicates a potential hazard situation or unsafe practice that, if not avoided,
could result in product or property damage.
Info: Provides application recommendations or other useful information to ensure that you
get the most from your product.
LF1031-INS-LAB-RevD12
3
Download from Www.Somanuals.com. All Manuals Search And Download.
2 SAFETY PRECAUTIONS
IMPORTANT: Before using patient lift, please read and adhere to the following safety precautions
and warnings. Failure to do so could result in serious personal injury or damage to your patient
lift.
Always consult your healthcare professional to determine safe methods most suitable for your
individual abilities. Protect yourself, your attendant and patient lift by having it serviced regularly.
If you experience any malfunction, contact your Lumex dealer immediately, as a hazardous
condition could result, causing personal injury or damage to your patient lift.
Periodic inspection, adjustment and replacement of worn parts are necessary to provide years of
excellent service. Refer to CARE AND MAINTENANCE section of this manual.
WARNINGS
WARNING: Important! Read and understand these instructions before assembling or
using the LF1031 and LF1041. If you do not understand any part of these warnings,
cautions or instructions, contact a healthcare professional for direction in the use of
this product. If the Lumex Patient Lift is not properly assembled, personal injury and
damage to the Lumex Patient Lift could result.
WARNING: If components are damaged or missing, contact your dealer immediately. DO
NOT use substitute parts. Use only Lumex replacement parts. The use of non-Lumex
replacement parts could cause personal injury, property damage, and void the warranty.
WARNING: The LF1031 Hydraulic Patient Lift / LF1041 Battery Powered Patient Lift
maximum weight capacity is 400 pounds (181 kg), EVENLY DISTRIBUTED. DO NOT
exceed the maximum weight capacity.
WARNING: GF Health Products, Inc. assumes no responsibility for any damage or
personal injury caused by improper assembly or use of this product.
WARNING: Check all parts for shipping damage before using. In case of damage, DO
NOT USE the equipment. Contact the carrier or your authorized Lumex dealer for further
instructions.
WARNING: Patient lift may tip over if used incorrectly. Read and adhere to the operating
instructions prior to lifting anyone.
WARNING: GF Health Products, Inc. strongly recommends that two caregivers take part
in the lifting process.
WARNING: When transferring a patient, always keep the patient centered over the base.
WARNING: This patient lift is designed for transfer only. It is not intended to be used for
transporting or moving a patient from one location to another.
WARNING: Never use the patient lift to lift or transfer anything other than a person.
WARNING: Ensure that lifting sling loops are correctly attached to the hooks to prevent
the patient from sliding or falling out of the sling, which could result in personal injury.
Ensure that sling grip handles are always positioned away from the patient so that the
smooth side is facing the patient.
WARNING: Do not lock or block the casters when lifting. The casters must be free to roll
so that the patient lift can stabilize as the patient is initially lifted from a chair, bed, or
any stationary object.
WARNING: To ensure stability while lifting or transferring a patient, the patient lift legs
must be locked in the maximum open position.
LF1031-INS-LAB-RevD12
4
Download from Www.Somanuals.com. All Manuals Search And Download.
WARNING: Replace any worn or broken parts immediately! Torn, cut, frayed, or broken
slings can fail, resulting in personal injury. Use only slings that are in good condition.
WARNING: Unauthorized modification of your patient lift or the use of non-Lumex
replacement parts may change the structure of the lift and could create a hazardous
condition, which may result in serious personal injury and will void the warranty.
WARNING: LF1031 Hydraulic Patient Lift only: The hydraulic pump is sealed at the
factory. If service is required, the unit must be returned to the factory for repair. DO NOT
attempt to open the hydraulic pump or obtain local service; this will void the warranty
and may result in damage to the unit. Consult your Lumex Dealer for further information.
WARNING: LF1041 Battery Powered Patient Lift only: The electronics and actuator /
motor contain no serviceable components. If service is required, the unit must be
returned to the factory for repair. DO NOT attempt to open the electronics or actuator /
motor or obtain local service; this will void the warranty and may result in damage to
the unit. Consult your Lumex Dealer for further information.
WARNING: Use extreme care, caution, and common sense when utilizing a patient lift.
Special attention must be given to patients who are disoriented and cannot cooperate
while being lifted, such as comatose, agitated, spastic, or mentally confused persons.
Consult their healthcare professional for assistance in lifting these types of patients.
WARNING: Lumex specifically disclaims responsibility for any bodily injury or property
damage which may occur during any use which does not comply with federal, state or
local laws or ordinances.
LF1031-INS-LAB-RevD12
5
Download from Www.Somanuals.com. All Manuals Search And Download.
3 FEATURES
Battery
(LF1041 only)
Boom
Battery charger
(LF1041 only)
Actuator
(LF1041 only)
Control box
(LF1041 only)
6-point spreader bar
(suspension)
LF1041 Battery Powered
Patient Lift
Hydraulic pump
(LF1031 Hydraulic Patient Lift only)
See above left for LF1041
Battery Powered Patient Lift
components
Mast
Push handle
Base adjustment
handle
Locking
casters
Adjustable base
LF1031 Hydraulic Patient Lift
Swivel
casters
LF1041 Battery Powered Patient Lift
(unless otherwise specified,
callout applies to both lifts)
LF1031-INS-LAB-RevD12
6
Download from Www.Somanuals.com. All Manuals Search And Download.
4 HANDLING PROCEDURES
SHIPPING AND TRANSPORTATION INSTRUCTIONS
If the patient lift is to be reshipped by common carrier, it should be packed in the same carton/s
in which it was received (the LF1031 is shipped in one carton; the LF1041 motor may be packaged
separately from the frame).
Unpacking
1. Check for any obvious damage to the carton or its contents. If damage is evident, notify the
carrier or your authorized Lumex Dealer.
2. Remove all loose packing from the carton.
3. Carefully remove all the components from the carton.
Info: Unless the patient lift is to be used immediately, retain boxes, containers and packing
material for use in storing until use of patient lift is required.
Inspection
1. Examine the exterior of the patient lift for nicks, dents, scratches or other damage.
2. Inspect all components.
Storage
1. Store the repackaged patient lift in a dry area.
2. DO NOT place other objects on top of the repackaged patient lift.
LF1031-INS-LAB-RevD12
7
Download from Www.Somanuals.com. All Manuals Search And Download.
5 ASSEMBLY INSTRUCTIONS
The LF1031 patient lift is packed in one carton. The LF1041 motor may be in a separate carton.
Components are shown in illustration below and called out in list below illustration.
COMPONENTS
3
4
7
6
5
1
2
8
LF1041 *
LF1031 Hydraulic Patient Lift components
Item Qty Description
1
1
Mast/Boom/Pump assembly (LF1031)
1
Mast/Boom assembly (LF1041)
* Info: Actuator / Motor, Battery, Battery Charger, Control Box, Hand Control Pendant,
Power Cable, and mounting items are packaged separately on LF1041.
2
3
4
5
6
7
8
1
1
2
1
1
1
1
Base
Base adjustment handle
Push handle
Spreader bar to boom attachment bolt
Spreader bar to boom attachment rod
Spreader bar
Assembly wrench set
LF1031-INS-LAB-RevD12
8
Download from Www.Somanuals.com. All Manuals Search And Download.
ASSEMBLY
Base and mast assembly
1. Remove the locking pin from the mast sleeve located on the rear of
the base.
remove locking pin
2. Install the L-inserts in the base.
install L-inserts
3. With the boom pointed in the same direction as the base legs, insert
the mast into the sleeve.
4. Insert the locking pin through the holes of the sleeve and mast.
install locking pin
5. Secure the locking pin by attaching the locking cable to the other
end of the locking pin.
secure locking pin
6. Screw the two mast attachment bolts into the holes in the sleeve of
the base.
7. Use the supplied hex wrench to tighten the bolts.
install mast bolts
LF1031-INS-LAB-RevD12
9
Download from Www.Somanuals.com. All Manuals Search And Download.
8. Screw the push handles into the mast and
tighten them.
install push handles
9. Insert spreader bar attachment rod into spreader
bar. Insert spreader bar rod into the U-bracket at
the end of the boom. Align holes. Insert spreader
bar attachment bolt and tighten.
install spreader bar
10. Install the base adjustment handle in the socket
located behind the mast sleeve on the base. With
base handle to the far right position, insert and
tighten base attachment bolt to ensure that the
handle is firmly attached to the base.
install base adjustment handle
11. LF1031 only: For shipping purposes, the boom is
in its lowest position. To elevate, turn the pump
knob clockwise until hand tight and operate the
pump handle.
The LF1031 lift is now ready for use.
LF1041 lift: continue to next page.
elevate boom
LF1031-INS-LAB-RevD12
10
Download from Www.Somanuals.com. All Manuals Search And Download.
mast
power bracket
Phillips head screw
LF1041 only: attaching
actuator / motor,
power bracket,
boom
components
Tools needed: Phillips screwdriver,
M3 hex key
mounting
points
1. Attach actuator / motor to
boom and mast using the
pre-attached bolts as shown
at near right; note that
mast
mount
secure
power bracket
power bracket
attach
control box
socket head
cap screw
actuator / motor
illustration shows mounting
points after assembly; shown
to boom and mast
components will not yet be present.
2. Use Phillips screwdriver to remove the three Phillips
heads screws on the back of the mast. Position power
bracket on mast as shown at upper far right. Secure
power bracket to mast by installing the Phillips head
mast screws as shown at upper far right (top screw is
shown). Use Phillips screwdriver to tighten.
mount
secure
3. Mount control box on mast by sliding its lower end onto
power bracket lip, as shown at near right. Secure control
box by installing socket head cap screw at top of control
box, as shown at far right. Use M3 hex key to tighten.
control box
control box
socket head cap screw
4. Mount battery charger on mast by sliding it onto power
bracket lip so it rests on top of control box. Secure
battery charger by installing socket head cap screw at
top of battery charger, as shown at right. Use M3 hex key
to tighten.
secure
battery charger
battery
5. Mount battery to mast by sliding it onto top power
bracket lip so it rests on top of battery charger, and boss
on battery back fits into power bracket hole.
battery boss
power bracket hole
mount battery
6. The control box, battery charger, and battery are shown
correctly mounted on the mast at right.
battery
7. Continue to next page for electrical power connections.
battery
charger
control box
power
components
mounted
LF1031-INS-LAB-RevD12
11
Download from Www.Somanuals.com. All Manuals Search And Download.
hand control pendant
connector
LF1041 only: electrical power connections
actuator / motor
connector
The control box is the heart of the electrical system.
The battery charger and battery are plugged into the control
box via built-in stacking connectors, depending on battery
charger options.
control box
The hand control pendant and actuator / motor are plugged
into the control box via the lower control box connectors,
shown at right.
hand control
pendant plug
When standing behind the patient lift, the hand control
pendant connection is on the far left of the control box.
control box connectors
1. Ensure the marks on the control box and the hand
control pendant plug are aligned, then firmly insert the
hand control pendant plug as shown at top right.
2. Immediately to the right of the hand control pendant
connection is the actuator / motor connection. Insert the
actuator / motor plug firmly in its receptacle (this plug
has nothing to align).
power cord
connector
and plug
3. Plug the power cord into its connector on the battery
charger as shown at right.
plug in power cord
power cord
hook
4. Slide the power cord hook onto the battery charger as
shown at right.
battery
charger
attach power cord hook
5. Snap the power cord hook onto the battery charger as
shown at right.
snap on power cord hook
6. Coil the power cord on the cord hook as shown at right.
7. Before proceeding, ensure that all plugs are seated firmly
in the proper receptacles.
power cord
coil power cord on hook
LF1031-INS-LAB-RevD12
12
Download from Www.Somanuals.com. All Manuals Search And Download.
LF1041 only: charging the battery
The battery used on the LF1041 has an audible alarm that sounds when the battery charge has
been depleted by 50%. Charge the battery as soon as possible upon hearing this alarm.
There is also a light on the hand control pendant that illuminates when it is time to recharge the
battery.
Info: This patient lift uses a lead acid battery. To prolong the life of the battery, charge the
battery fully for 24 hours before initial use. To increase the longevity of the battery, keep the
battery fully charged when not in use. Subsequent battery charging time is approximately
four hours.
WARNING: After the alarm sounds, there could still be sufficient battery power left
to lift a patient. Be aware that a lifted patient cannot be lowered when the battery is
discharged. Refer to the Emergency Procedures section that follows for discharged
battery operation.
Charging the battery with lift-mounted battery charger: With control box, battery charger, and
battery assembled, insert the power cord plug into a standard 110V receptacle. A green light
signifying power and a yellow light signifying that the battery is charging will illuminate on the
battery charger. When the battery is fully charged, the yellow light will shut off. Unplug the power
cord from the receptacle and replace it on the power cord hook.
Info: The patient lift will not operate during lift-mounted battery charging.
The LF1041 lift is now ready for use.
Charging the battery with wall-mounted battery charger: Remove the battery from its bracket
on the patient lift by placing your hand on top of the battery, curling your fingers behind the
battery, and pulling up on the metal handle. Pull the battery up and toward you to remove. Place
the battery on top of the wall-mounted battery charger. A green light signifying power and a yellow
light signifying that the battery is charging will illuminate on the battery charger. When the battery
is fully charged, the yellow light will shut off. When battery charging is complete, place the battery
back on the patient lift. Ensure the battery tabs are fully engaged with the bracket.
The LF1041 lift is now ready for use.
power
switch
LF1041 only: control box power switch
(emergency stop button)
In order to use the lift, the control box power switch must be turned
ON. To turn the control box power switch ON, turn the red switch
button clockwise, it will pop out to the ON position.
To turn the control box power switch OFF, hit it firmly with the palm
of your hand; the switch will lock in the depressed position, stopping
all motor function.
To prolong the life of the battery, turn the control box power switch
OFF when the lift is not in use.
control box power switch
LF1031-INS-LAB-RevD12
13
Download from Www.Somanuals.com. All Manuals Search And Download.
6 OPERATING INSTRUCTIONS
LF1041 ONLY: EMERGENCY PROCEDURES
emergency
stop button
Emergency stop: The control box is equipped with a red
emergency stop button. This button, also known as the
control box power switch, is to be used in the event the hand
control pendant ceases to function correctly. To engage, hit
the button firmly with the palm of your hand. The button will
lock in the depressed position, stopping all motor function.
To disengage, turn the button clockwise; it will pop out and
return to the ON position.
Emergency lowering: The motor is equipped with an
emergency release, which is located on top of the motor
as shown in the picture at right. To engage, pull up on the
release, which will disengage the gearing in the cylinder, and
lower the patient. The emergency release is adjustable by
turning the screw adjacent to the pull stem. Turn the screw
clockwise to increase and counter-clockwise to decrease the
sensitivity of the release mechanism.
emergency stop button
motor
adjustment
screw
emergency
release
WARNING: Use caution when using the emergency
release. Pull back slowly on the release, which will
gradually lower the patient. Pulling back too quickly
could cause the patient to be lowered too rapidly,
which could result in serious personal injury. Do
not, under any circumstances, use the emergency
release for lowering the patient when the hand control
pendant and battery are operating correctly. The
emergency release should not be used in lieu of the
hand control pendant and battery.
emergency release
LF1031-INS-LAB-RevD12
14
Download from Www.Somanuals.com. All Manuals Search And Download.
LIFT OPERATION: PATIENT TRANSFER
WARNING: Do not attempt to transfer a patient without prior approval of the patient’s
healthcare professional and without having read the instructions and practiced using
the patient lift.
WARNING: Use special care for those patients who cannot cooperate while being lifted.
WARNING: Ensure all bolts, hooks, and loops are securely fastened.
WARNING: Use only Lumex slings with Lumex patient lifts. DO NOT exceed the maximum
weight capacity of the selected sling.
WARNING: DO NOT use torn, cut, frayed or broken slings as personal injury could result.
WARNING: Ensure the floor area is clear of any obstacle that could cause a caster to
stop and possibly tip the patient lift.
WARNING: GF Health Products, Inc. strongly recommends that two caregivers take part
in the lifting process.
Info: Instructional videos that review the application of several sling styles and lifting
Lifting the patient from a bed
1. With patient face-up in center of bed, roll patient onto their side away from attendant.
2. Place sling, folded lengthwise, halfway under patient with lower edge just beneath knees.
3. Roll patient onto their side toward attendant, pull sling to smooth it, and center patient on
sling.
4. Pull the leg loops forward and under the thighs.
5. Cross the loops, one through the other.
6. Roll the base of the patient lift under the bed, locating the spreader bar over the patient. Be
careful not to lower the spreader bar onto the patient. Center the spreader bar over the patient
before lifting.
WARNING: Do not lock or block the patient lift casters when lifting. The casters must be
free to roll so that the patient lift can stabilize as the patient is lifted from the bed.
WARNING: To ensure stability while lifting or transferring a patient, the patient lift legs
must be locked in the maximum open position.
7. Hook the sling loops onto the spreader bar.
8. If the patient is in a hospital bed, raise the bed's head section to elevate the patient.
9. Raise the patient carefully. If possible, lower the bed. Reposition the patient as needed, using
the handle on the back of the sling, to ensure a safe and comfortable sitting position as the
patient is raised.
10. Raise the patient until the buttocks are just above the mattress. Lift the patient’s legs and turn
the patient to face the lifting column. Move the patient lift away from the bed. Ensure that the
patient's arms and legs clear the patient lift, bed, or any other obstacles.
Lowering the patient onto a bed
1. Raise the patient lift until the patient’s buttocks are above the surface of the bed. Ensure that
there are no obstacles underneath or around the bed. Move the patient lift under the bed.
Ensure the patient does not hit any portion of the bed and that the patient’s legs have cleared
the bed.
LF1031-INS-LAB-RevD12
15
Download from Www.Somanuals.com. All Manuals Search And Download.
2. Center the patient over the bed, and rotate the patient to face the foot section of the bed.
SLOWLY release the pressure knob by turning it counter-clockwise for gradual lowering of the
patient (LF1031) or using the hand control pendant (LF1041). Carefully lower the patient onto
the bed. Support the patient’s head, if necessary, while being lowered. Remove the sling by
reversing the procedures used when lifting patient from bed.
Lifting the patient from the floor
1. Position the patient lift behind the patient, placing a pillow over the base to protect and
support the patient’s head and neck. Place the patient's head carefully on the base. The patient
lift should be positioned with the spreader bar centered over the patient.
2. Lower the spreader bar. Attach the sling to the spreader bar hooks.
3. Lift the patient up above the floor. Position the patient in the sling by pulling the handle on the
back of the sling. Turn the patient to face the lifting column, and place the patient’s feet on the
base, straddling the column for extra stability.
Lowering the patient to the floor
1. Place a pillow on the base to protect the patient’s head and neck (this is unnecessary when
using a sling with a head support).
2. Reverse the above procedures to lower.
Transferring the patient from a chair to a wheelchair or commode chair
Info: GF Health Products, Inc. recommends that two attendants be used when transferring a
patient to and from a wheelchair.
1. Fold the sling.
2. While maintaining support, gently lean the patient forward. Place the folded sling behind the
patient’s back and push it down until it touches the seat of the chair.
3. Use the loops to pull the leg sections of the sling forward and beneath the patient’s thighs.
Then cross the loops, one through the other.
4. Move the patient lift around to the front of the chair.
5. Hook all sling loops onto the spreader bar.
6. Carefully raise the patient.
7. The second attendant should now move the wheelchair into position.
8. Engage the wheelchair wheel locks to prevent movement.
WARNING: Before transfer, ensure wheelchair wheel locks are in locked position.
Wheelchair wheel locks must be in a LOCKED POSITION before lowering the patient into
the wheelchair, or unexpected wheelchair movement could result, which may result in
serious personal injury to both the patient and the attendants.
9. Position the patient over the wheelchair with their back against the back of the wheelchair.
10. SLOWLY lower the patient into the wheelchair.
11. With one attendant behind the wheelchair and the other operating the lift, the attendant
behind the chair should pull back on the handle or sides of the sling to place the patient into
the back of the chair. This will maintain a good center of balance and prevent the chair from
tipping forward.
LF1031-INS-LAB-RevD12
16
Download from Www.Somanuals.com. All Manuals Search And Download.
Removing the sling
Gently pull the leg sections to the side, out from under the patient’s thighs. Standing by the
patient’s side, lean the patient forward while supporting the patient with one hand. Pull the sling
up from behind the patient’s back and position the patient in the chair.
WARNING: Pulling sharply on the sling may cause the patient to fall forward, resulting
in personal injury. Always support the patient by keeping one arm around the patient’s
shoulders when placing or removing the sling.
Info: If possible, the patient should be encouraged to take an active part in removing the
sling. The patient can lean forward, lift their thighs, and help in removing the leg sections.
Transferring the patient from a bed to a wheelchair or commode chair
1. For smooth, easy lifting, have the wheelchair or commode chair in ready position with wheel
locks in locked position.
2. With patient face-up in center of bed, roll patient onto their side away from attendant.
3. Place sling, folded lengthwise, halfway under patient with lower edge just beneath knees.
4. Roll patient to side toward attendant, pull sling to smooth it, and center patient on sling.
5. Pull the leg loops forward and under the thighs.
6. Cross the loops, one through the other.
7. With base of lift under bed, SLOWLY release pressure knob by turning it counter-clockwise and
pulling the boom down (LF1031) or using the hand control pendant (LF1041). Attach sling to
spreader bar.
8. Turn pressure knob clockwise until tight, and lift the patient by operating the pump handle.
9. Lift patient, maintaining patient's position facing attendant, until patient's feet swing easily off
the bed.
10. Position patient above wheelchair or commode chair. Lock lift brakes.
WARNING: Before transfer, ensure wheelchair or commode chair wheel locks are in
locked position. Wheelchair wheel locks must be in a LOCKED POSITION before lowering
the patient into the wheelchair, or unexpected wheelchair movement could result, which
may result in serious personal injury to both the patient and the attendants.
11. SLOWLY release pressure knob by turning it counter-clockwise and pulling the boom down
(LF1031) or using the hand control pendant (LF1041) to gradually lower patient. During
descent, assist patient in attaining correct sitting posture.
12. When patient is seated, push down on boom to slacken sling loops. Patient can remain seated
on sling.
13. For commode chair use, adjust clothing before moving lift to straddle commode chair. Keep
sling loops taut and ensure that patient is in a comfortable position.
LF1031-INS-LAB-RevD12
17
Download from Www.Somanuals.com. All Manuals Search And Download.
7 CARE AND MAINTENANCE
Proper care and maintenance are essential to keeping your Lumex patient lift in a safe operating
condition. In addition to inspecting the unit before each use, periodic maintenance checks should
be done. We further recommend that this unit be serviced at least once annually by an authorized
dealer.
WARNING: It is extremely important that the patient lift be inspected before each use.
Ensure that all hardware and accessories are secure and that the actuator / motor is
functioning properly. Failure to do so could result in patient / attendant personal injury
or damage to your patient lift. Torn, cut, frayed or broken slings can fail, resulting in
serious personal injury. Only use slings in good condition. Inspect before each use.
Destroy and discard old, worn and unusable slings.
When you believe a component or part is not functioning properly, immediately contact your
authorized Lumex dealer, as a potentially hazardous condition could exist.
BEFORE EACH USE
Before each use of the patient lift, ensure that:
• All nuts and bolts are tight
• The lift moves freely (with caster brakes off)
• Rear caster brakes operate properly
• Base adjustment mechanism operates easily
• Spreader bar clips are secured tightly
• The hydraulic pump operates properly (LF1031 only)
The electronic components operate properly (LF1041 only)
BATTERY
To increase the longevity of the battery, plug in the charger whenever possible (LF1041 only).
LUBRICATION
Your Lumex patient lift was designed with minimum maintenance in mind; however, a six (6)
month check and lubrication should ensure that your patient lift will remain in peak operating
condition. The casters MUST roll and swivel freely. Lightweight waterproof grease (such as
automobile lubricant) should be applied to the ball bearings of the casters every six (6) months.
Apply grease more frequently if the casters are exposed to periods of excess moisture. Lubricate
all pivot points with lightweight oil. Clean patient lift of all excess oil and grease.
Before use of the patient lift, and every two months thereafter, place a drop of oil on the following
areas:
• The point where mast and boom connect
• Spreader bar hooks
• Pump handle hinge
• Caster axles
• The point where pump and boom connect (LF1031 only)
The point where actuator / motor and boom connect (LF1041 only)
LF1031-INS-LAB-RevD12
18
Download from Www.Somanuals.com. All Manuals Search And Download.
CLEANING
Regular cleaning is recommended. A soft cloth, dampened with water and a mild NON-ABRASIVE
detergent (household cleaner, soapy water, or hospital grade disinfectant) is all that is needed to
clean your patient lift. Automobile wax or furniture polish will help maintain the finish over a long
period of time.
s NOTICE: DO NOT wash under water pressure or steam clean.
CHECKING FOR DAMAGE AND WEAR
It is extremely important to frequently check the areas of the patient lift that are under a great deal
of stress for damage and wear. These areas include slings, spreader bars, spreader bar suspension
and mounting hardware, pump mounting hardware and all pivot points. Inspect for signs of
cracking, fraying, deformation, and any loose components
WARNING: Replace any defective parts immediately, and make certain that the patient
lift is not used until repairs are made. Contact your authorized Lumex dealer for
replacement parts.
CARE OF SLINGS
Lumex slings are washable at 150° F and autoclavable at 185° F for 30 minutes. The head support's
plastic insert should be removed before washing. The slings can be dried by using a dryer on LOW
heat or by hanging.
s NOTICE: DO NOT bleach slings!
LF1031-INS-LAB-RevD12
19
Download from Www.Somanuals.com. All Manuals Search And Download.
LF1031 ONLY: INSTALLATION OF 1010S005 HYDRAULIC PUMP
Removal of Existing Pump
upper boom
bracket
lower mast
bracket
boom
existing pump
mast
existing pump
detach existing pump
from upper boom
detach existing pump
from lower mast
1. Follow the instructions in this section to remove the existing pump, shown at above left. If
installing pump for the first time, proceed to Installation of new pump on next page.
2. Detach the top of the existing pump from the upper boom bracket, as shown at above center:
9
Use two /16" wrenches to remove the hardware that secures top of pump to upper boom
bracket. This will free the pump to rotate downward.
3. Detach the bottom of the existing pump from the lower mast bracket, as shown at above right.
9
Use two /16" wrenches to remove the hardware that secures bottom of pump to lower mast
bracket. This will free the pump — take care that it does not fall.
LF1031-INS-LAB-RevD12
20
Download from Www.Somanuals.com. All Manuals Search And Download.
Installation of New Pump
9
Tools needed: two /16" wrenches
Info: The hardware to attach the Hydraulic Pump (GF #1010S005) to the LF1031 mast and
boom comes with the lift frame.
boom
mast
upper boom
bracket
attach new pump
to lower mast
detail — attach new
pump to lower mast
attach new pump
to upper boom
1. Attach the bottom of the new hydraulic pump to the lower mast bracket with the pressure
knob and pump lever toward the bottom of the pump, as shown at above left, and the “FRONT”
label toward you, as shown at above center. Use a hex bolt, washers, and hex nut to secure
9
pump to mast. Use two /16" wrenches to tighten.
2. Attach the top of the hydraulic pump to the upper boom bracket, as shown at above right. Use
9
a hex bolt, washers, and hex nut to secure pump to boom. Use two /16" wrenches to tighten.
WARNING: Ensure that the Hydraulic Pump is installed as described. If the Hydraulic
Pump is not properly installed, personal injury and damage to the Hydraulic Pump could
result.
LF1031-INS-LAB-RevD12
21
Download from Www.Somanuals.com. All Manuals Search And Download.
MAINTENANCE SCHEDULE
Info: Only qualified persons should service and repair your Lumex patient lift. Regular
maintenance of your patient lift is necessary to ensure continuing proper and safe
operation. Read and observe the following recommended maintenance schedule:
Item
Inspect
initially
Inspect/adjust
daily
monthly every 6
months
Base and Casters
Ensure that base is level (all four wheels touch the floor)
Ensure that caster mounting bolts and axle bolts are tight
Ensure that casters swivel and roll freely
Apply a light grease to caster ball bearings
Base Adjustments
Ensure that base opens and closes smoothly
Ensure that base locks securely in three positions
Ensure that locking knob is in place and operating properly
Inspect all base hardware for wear
Mast
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Ensure that mast is fully engaged and locked into mast sleeve
Inspect mast for bends or deflections
Inspect all mast hardware for wear
Boom
Inspect boom for bends or deflections
Ensure that boom is centered between the base legs
Inspect all boom hardware for wear
Spreader Bar
X
X
X
X
X
X
X
X
X
X
X
Inspect hanger post for wear or damage
Inspect sling hook for wear or deflections
Inspect all spreader bar hardware for wear
Slings
X
X
X
X
X
Inspect slings and straps for wear or damage
Inspect metal/plastic parts for wear or damage
Lubrication with light oil
X
X
Lubricate all pivot points
X
X
Lubricate top and bottom of actuator / motor mounting
Cleaning
Clean with mild non-abrasive detergent
X
Pump (LF1031 only)
Ensure that pump mounting hardware is tight
Inspect pump mounting hardware for wear
Electronics and actuator / motor (LF1041 only)
Ensure that battery is fully charged
Inspect for proper up / down operation
Ensure that mounting hardware is tight
Inspect mounting hardware for wear
X
X
X
X
X
X
X
X
X
X
LF1031-INS-LAB-RevD12
22
Download from Www.Somanuals.com. All Manuals Search And Download.
8 TROUBLESHOOTING
WARNING: If you experience a problem with your lift and are unable to service it
yourself, contact your Lumex dealer.
LF1031 HYDRAULIC PATIENT LIFT
SYMPTOM
POSSIBLE CAUSE
ACTION
Patient lift does not go up
or down
Air in the cylinder
Bleed the cylinder by turning the release
knob and push down on the mast. Pump
up the lift again and turn the release knob
again.
Loss of hydraulic fluid (leaking
cylinder)
Replace any leaking cylinder
Patient lift does not go
down
Air in the cylinder
See above Directions
Info: If the hydraulic cylinder begins leaking fluid, do not operate lift. Contact your Lumex
dealer for a new cylinder.
LF1041 BATTERY POWERED PATIENT LIFT
SYMPTOM
POSSIBLE CAUSE
ACTION
Control box is not
responding and relay is not
clicking
Battery discharged
Charge battery
Battery charger defective
Battery defective
Replace battery charger
Replace battery
Hand control pendant defective
Actuator plug not properly seated
Actuator defective
Replace hand control pendant
Push actuator plug firmly into control box
Replace actuator
Actuator / motor does not
run, but relay is clicking
Control box defective
Replace control box
Actuator / motor goes up
or down, but not both
Hand control pendant defective
Control box defective
Replace hand control pendant
Replace control box
LF1031-INS-LAB-RevD12
23
Download from Www.Somanuals.com. All Manuals Search And Download.
9 SPECIFICATIONS
Info: All dimensions are given in inches unless otherwise specified.
LF1031 HYDRAULIC PATIENT LIFT / LF1041 BATTERY POWERED PATIENT LIFT
Base width
Open: 39.4"
Closed: 25.6"
Base height clearance
Base length
5-1/2"
45.30"
Lifting height range Info: As measured from point
where spreader bar attaches to boom.
Minimum: 27.4"
Maximum: 77.7"
Range: 50.3"
Mast height
55.1"
Weight capacity
Caster size
400 lb (181 kg), EVENLY DISTRIBUTED
4.00" diameter, both locking and non-locking
Product weight
Shipping weight
90 lb
103.2 lb
LF1031-INS-LAB-RevD12
24
Download from Www.Somanuals.com. All Manuals Search And Download.
10 LIMITED WARRANTY
GF Health Products Inc. warrants this product to be free from manufacturing defects in material
and workmanship for its period of warranty that is in accordance with industry standards. This
warranty is extended only to the original purchaser/consumer or dealer/non-consumer of this new
product and to no other purchaser or transferee.
GF Health Products, Inc. warrants the Lumex® LF1031 Hydraulic Patient Lift and its components to
be free from defects in workmanship and materials as listed below:
Frame / Parts: 1 Year
Hydraulic Pump: 1 Year
GF Health Products, Inc. warrants the Lumex® LF1041 Battery Powered Patient Lift and its
components to be free from defects in workmanship and materials as listed below:
Frame / Parts: 1 Year
Electronics: 1 Year
Actuator / motor: 1 Year
The Warranty period for the consumer commences on the first date a product is delivered to
consumer by seller/dealer. If the product is rented or leased, the warranty period commences
on the invoice date from GF Health Products, Inc. A copy of the invoice showing date of purchase
must be provided when submitting warranty claims. When proof of purchase date is not provided,
warranty coverage shall commence upon GF Health Products, Inc.’s invoice date to the dealer/
purchaser.
If within the warranty period, the product or component part is proven to GF Health Products, Inc.’s
satisfaction to be defective, GF Health Products, Inc. shall provide, at its option, one of the following:
(1) repair or replacement of any defective or nonconforming part or product or (2) a credit and/
or refund of the original selling price. GF HEALTH PRODUCTS, INC.’S SOLE OBLIGATION AND YOUR
EXCLUSIVE REMEDY UNDER THIS WARRANTY SHALL BE LIMITED TO SUCH REPAIR, REPLACEMENT,
CREDIT AND/OR REFUND. This warranty does not include any labor charges incurred in replacement
part(s) installation or any associated freight or shipping charges to the manufacturer.
For warranty service, please contact the authorized distributor from whom you acquired your GF
Health Products, Inc. product. Upon receiving notice of an alleged defect in a product,GF Health Products,
Inc. will issue a return authorization. The defective product or part(s) must then be returned, at the
purchaser’s cost, for warranty inspection using the serial number as identification (or, if the product
is not serialized, lot number and date code) within thirty (30) days of return authorization issue date.
In the event you do not receive satisfactory warranty service, please contact GF Health Products, Inc.
at the address below. DO NOT return products to our factory without prior authorization.
LIMITATIONS AND EXCLUSIONS: The foregoing warranty excludes consumables, such as filters
and tubing, as these require routine maintenance. The foregoing warranty shall not apply to serial
numbered products if the serial number has been removed or defaced. Products subjected to
negligence, abuse, misuse, improper operation, improper maintenance, improper cleaning, improper
storage, or damages beyond GF Health Products, Inc.’s control are not covered by this warranty, and
that evaluation shall be solely determined by GF Health Products, Inc. This warranty shall not apply
to problems arising from normal wear and tear or failure to follow instructions. The warranty shall
also not apply to products modified without GF Health Products, Inc.’s express written consent; nor
shall it apply if parts not manufactured by GF Health Products, Inc., or if parts not complying with
original equipment specifications are added to GF Health Products, Inc. products, or if the product or
part is serviced by an entity not authorized by GF Health Products, Inc.
LF1031-INS-LAB-RevD12
25
Download from Www.Somanuals.com. All Manuals Search And Download.
THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS
WARRANTIES AND IMPLIED WARRANTIES, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE, AND SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS
WARRANTY PROVIDED HEREIN, AND THE REMEDY FOR VIOLATIONS OF ANY IMPLIED
WARRANTY SHALL BE LIMITED TO THE REPAIR, REPLACEMENT, CREDIT AND/OR
REFUND OF THE DEFECTIVE PRODUCT OR PART PURSUANT TO THE TERMS CONTAINED
HEREIN. GF HEALTH PRODUCTS, INC. SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL
OR INCIDENTAL DAMAGES WHATSOEVER.
This warranty gives you specific legal rights and you may also have other legal rights which vary
from state to state (province to province). Some states (provinces) do not allow the exclusion or
limitation of incidental or consequential damage, or limitation on how long an implied warranty
lasts, so the above exclusion and limitations may not apply to you.
The warranties contained herein contain all the representations and warranties with respect
to the subject matter of this document, and supersede all prior negotiations, agreements and
understandings with respect thereto. The recipient of this document hereby acknowledges and
represents that it has not relied on any representation, assertion, guarantee, warranty, collateral
contract or other assurance, except those set out in this document.
GF Health Products, Inc. (“Graham-Field”)
2935 Northeast Parkway
Atlanta, GA 30360
Tel 770-368-4700, Fax 770-368-2386
LF1031-INS-LAB-RevD12
26
Download from Www.Somanuals.com. All Manuals Search And Download.
11 INDEX
S
A
Safety precautions 4
Safety statements, significance of 3
Shipping and transportation instructions 7
Specifications 24
Assembly instructions 8
B
Battery care 18
Before each use 18
Storage 7
C
T
Care and maintenance 18
Transferring the patient from a bed to a wheelchair or
commode chair 17
Care of slings 19
CAUTION statement, significance 3
Checking for damage and wear 19
Cleaning 19
Transferring the patient from a chair to a wheelchair or
commode chair 16
Troubleshooting 23
Components 8
U
F
Unpacking 7
Features 6
V
H
Handling Procedures 7
W
I
Warnings 4
Info statement, significance 3
WARNING statement, significance 3
Warranty, limited 25
Weight capacity 4
Inspection 7
Introduction 3
L
LF1031 only: hydraulic pump 1010S005, installation 20
LF1041 only: attaching power bracket and components
11
LF1041 only: charging the battery 13
LF1041 only: electrical connections 12
LF1041 only: emergency procedures 14
Lifting the patient from a bed 15
Lifting the patient from the floor 16
Lift operation: patient transfer 15
Lowering the patient onto a bed 15
Lowering the patient to the floor 16
Lubrication 18
M
Maintenance schedule 22
N
NOTICE statement, significance 3
O
Operating instructions 14
R
Removing the sling 17
LF1031-INS-LAB-RevD12
27
Download from Www.Somanuals.com. All Manuals Search And Download.
GF Health Products, Inc.
2935 Northeast Parkway
Atlanta, Georgia 30360 USA
tel: 770-368-4700
Made in Taiwan
Download from Www.Somanuals.com. All Manuals Search And Download.
|