750000
ALTERNATING PRESSURE /
LOW AIR LOSS MATTRESS SYSTEM
USER MANUAL
Important: Do not operate the Mattress System
without first reading and understanding this manual!
Save this manual for future use.
Info: The most current version of this manual can be
750000-INS-LAB-RevC11
Download from Www.Somanuals.com. All Manuals Search And Download.
INTRODUCTION
Use this manual for set-up and operation of the Lumex
AltaDyne 750000 Alternating Pressure / Low Air Loss
Mattress System. Read all instructions before using the
mattress system. Save this manual for future reference.
INTENDED USE OF THIS DEVICE
The intended use of the Lumex AltaDyne 750000
Alternating Pressure / Low Air Loss Mattress System is:
• The treatment of pressure ulcers while optimizing user
comfort.
• Pain management as prescribed by a physician.
• As described above, in either a homecare or long term
care setting.
Contraindication
WARNING: DO NOT use this product in the presence
of flammable anesthesia. There is a possible fire
hazard when this product is used with certain oxygen
delivery equipment.
1
Use nasal cannula, face mask, or /2 length oxygen
tent to deliver oxygen in the presence of this product.
DO NOT use a full length oxygen tent that extends
past the top surface of the mattress with this product.
750000-INS-LAB-RevC11
3
Download from Www.Somanuals.com. All Manuals Search And Download.
IMPORTANT SAFETY PRECAUTIONS
The safety statements presented in this chapter refer
to the basic safety information that should be observed
by those using the Alternating Pressure / Low Air Loss
Mattress System. There are additional safety statements in
other chapters or sections, which may be the same as or
similar to the following, or specific to the operations.
DANGER: Indicates an imminent hazard situation
that, if not avoided, will result in death or serious
injury.
WARNING: Indicates a potential hazard situation or
unsafe practice that, if not avoided, could result in
death or serious injury.
CAUTION: Indicates a potential hazard or unsafe
practice that, if not avoided, could result in minor or
moderate personal injury.
s NOTICE: Indicates a potential hazard or unsafe
practice that, if not avoided, could result in product
or property damage.
Info: Provides application recommendations or other useful
information to ensure that you get the most from your
product.
DANGER
To reduce the risk of electrocution:
DANGER: Always unplug this product immediately
after using.
DANGER: Do not use this product while bathing.
DANGER: Do not place or store this product where it
can fall or be pulled into a tub or sink.
DANGER: Do not place this product in or drop into
water or other liquid.
DANGER: Do not reach for a product that has fallen
into water. Unplug it immediately.
4
750000-INS-LAB-RevC11
Download from Www.Somanuals.com. All Manuals Search And Download.
WARNING
To reduce the risk of burns, electrocution, fire, or
personal injury:
WARNING: Do not leave this product unattended
when plugged in.
WARNING: Always use close supervision when this
product is used by, on, or near children or those who
require close supervision.
WARNING: Use this product only as intended and
described in this manual. Do not use attachments or
accessories not recommended by Graham-Field.
WARNING: Never operate this product if:
a) It has a damaged cord or plug.
b) It is not working properly.
c) It has been dropped or damaged.
d) It has been dropped into water.
Return the product to your Graham-Field equipment
provider.
WARNING: Keep the cord away from heated surfaces.
WARNING: Never drop or insert any object into any
opening or hose.
WARNING: Never block the air openings of this
product or place it on a soft surface, such as a bed or
couch, where the openings may be blocked. Keep the
air opening free of lint, hair, and other similar particles.
WARNING: Connect this product only to a properly
grounded power outlet.
WARNING: Maximum patient weight capacity for this
product is 350 lb (159 kg).
750000-INS-LAB-RevC11
5
Download from Www.Somanuals.com. All Manuals Search And Download.
WARNING: Patient entrapment with bed side rails
may cause injury or death. The bed frame and its
components, including the mattress, bed side rails,
head and foot board, bedding, and any accessories
added to the bed, can all affect the risk of entrapment.
Thorough patient assessment and monitoring are
necessary to reduce the risk of entrapment, including
establishing whether the use of a bed rail is in the
best interest of the patient. Read and understand the
User Manual before using this equipment. GF Health
Products, Inc. product manuals are available online at
learn more about the risks of entrapment. It is the
responsibility of the facility and provider to be in
compliance with these guidelines. Refer to user
manuals for beds and rails for additional product
safety information.
After any adjustment, repair or service, and before
use, ensure all attaching hardware is securely
tightened. Bed rails with dimensions different
than the original equipment specified by the bed
manufacturer may not be interchangeable and may
result in entrapment or other injury.
NOTICE
s NOTICE: The pump can be used only with the
accompanying mattress. Do not use it for any other
purpose.
6
750000-INS-LAB-RevC11
Download from Www.Somanuals.com. All Manuals Search And Download.
EQUIPMENT SYMBOLS
Attention! Consult accompanying documents (this manual)
“BF”symbol: Indicates that this product is in accordance
with the degree of protection against electric shock for type
BF equipment
Functional earth (for UL only)
Class II (for UL only)
Disposal of Electrical & Electronic Equipment (WEEE):
Do not treat this product as household waste. For more
detailed information with regard to returning and recycling
this product, please consult your local city office, household
waste disposal service, or Graham-Field equipment provider.
For U.S. and Canada only
Medical Equipment: Air Pump
E228589
53DG
with respect to electrical shock, fire and
mechanical hazards only in accordance with
UL60601-1 and CAN/CSA C22.2 No. 601.1
750000-INS-LAB-RevC11
7
Download from Www.Somanuals.com. All Manuals Search And Download.
PRODUCT DESCRIPTION
PUMP AND MATTRESS SYSTEM
foot end
symbol
CPR valves
mattress
pump
mattress-pump
connectors
pump and mattress system
Pump: front
quick connector
8
750000-INS-LAB-RevC11
Download from Www.Somanuals.com. All Manuals Search And Download.
Pump: rear
Pump: top panel
static function switch
main power switch
low pressure indicator
normal pressure indicator
pressure adjustment knob
Pressure adjust knob: The pressure adjust knob controls
the air pressure output. Turn it clockwise to increase
output pressure (firmer); turn it counter-clockwise to
decrease air output pressure (softer). Higher pressure
output will better support the heavier-weight user. The
setting is measured in kilograms (kg). Please consult your
physician for an appropriate setting.
Normal pressure indicator: The green normal pressure LED
illuminates when the pressure inside the air mattress has
reached the selected pressure setting.
Low pressure indicator: The yellow low pressure LED
illuminates when the pressure inside the air mattress is
below the selected setting.
750000-INS-LAB-RevC11
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Static function switch: Press the static function switch to
suspend alternating function, if needed; the pressure inside
the air tubes will then be adjusted to identical pressure.
Press the static function switch again to change back
to alternating mode. The static function switch toggles
between static and alternating modes.
Main power switch: The main power switch toggles
between ON and OFF. When power is ON, either the normal
or low pressure indicator illuminates.
10
750000-INS-LAB-RevC11
Download from Www.Somanuals.com. All Manuals Search And Download.
INSTALLATION
UNPACKING
1. Before unpacking, check for obvious damage to the
carton or its contents. If damage is evident, please
notify the carrier and your Graham-Field equipment
provider.
2. Remove all loose packing from the carton.
3. Carefully remove all components from the carton.
4. Inspect all components. If damage is evident, please
notify the carrier and your Graham-Field equipment
provider.
750000-INS-LAB-RevC11
11
Download from Www.Somanuals.com. All Manuals Search And Download.
PUMP AND MATTRESS INSTALLATION
1. Place the mattress on top
of the bed frame; ensure
that the foot end symbol
faces up at foot end as
shown at right.
2. Fold open the wire
hangers on the back of
the pump as shown at
right.
3. Either hang the pump
on the bed's foot end
as shown at right, or
place the pump on
a flat surface easily
accessible to the user,
caregiver, and/or
doctor.
s NOTICE: Place the device in a position where the
user, caregiver, and/or doctor can access it easily.
4. Connect the mattress-
pump air hose quick
connector; ensure that
the connector has clicked
securely into place before
continuing.
s NOTICE: Ensure that the air hoses are not kinked or
tucked under the mattress.
12
750000-INS-LAB-RevC11
Download from Www.Somanuals.com. All Manuals Search And Download.
WARNING: Locate all cords so that they will not be
stepped on, tripped over, or otherwise subjected to
damage or stress.
WARNING: Do not use a household extension cord if
the electrical cord does not reach the power outlet.
Use of an improper extension cord could result in
fire and electric shock. If an extension cord must
be used, use a three conductor cord with ground,
properly wired, in good electrical condition, and keep
it as short as possible.
WARNING: Ensure that the local power voltage is
appropriate for the pump unit.
5. Plug the power cord into a properly grounded electrical
outlet.
750000-INS-LAB-RevC11
13
Download from Www.Somanuals.com. All Manuals Search And Download.
OPERATION
1. Turn the power switch
to the ON position.
2. Adjust the pressure
adjust knob to the
maximum position.
Info: Set the pressure adjust knob to the maximum position
for the quickest inflation at set-up; the mattress can then be
adjusted to the desired pressure.
3. The pump will begin
delivering air into the
mattress. While the
mattress is filling, the
low pressure indicator (yellow LED) will illuminate.
4. When the selected
pressure level is
reached, the low
pressure indicator will
de-illuminate, and the normal pressure indicator (green
LED) will automatically illuminate.
5. Replace all sheets, blankets, and pillows on the bed.
PRESSURE ADJUSTMENT
Generally, a lighter patient will need a lower (softer)
setting while a heavier patient will need a higher (firmer)
setting, but pressure adjustment must ultimately be based
on the patient's weight distribution. A short, stocky person
may be the same weight as a tall, thin person, but their
settings will not be the same. The mattress should be firm
enough to support the patient, yet soft enough so the
patient will sink down to eliminate pressure points. Test
the pressure by placing four fingers below the mattress
14
750000-INS-LAB-RevC11
Download from Www.Somanuals.com. All Manuals Search And Download.
under the patient's buttock area; your fingers should just
fit. If the fit is too loose, decrease the setting. If your fingers
do not fit well, increase the setting two notches firmer. Wait
ten minutes and then check again. You should only have
to adjust this setting one time per patient. Please consult
your physician for an appropriate setting.
QUICK DISCONNECT
Pull the power plug from the wall connector to disconnect
the device quickly.
quick connector
TRANSPORT CAP
In case of power failure or transport:
Disconnect the quick connector and
cap the mattress with the transport
cap to slow deflation.
CPR FUNCTION
transport cap
If CPR must be performed on the
patient while the mattress is in use,
to immediately deflate the mattress,
either:
1. Pull the CPR valves from the
mattress in the location of the
arrow shown at right, or
2. Disconnect the quick connector
from the pump in the location of
the arrow shown at right.
750000-INS-LAB-RevC11
15
Download from Www.Somanuals.com. All Manuals Search And Download.
MAINTENANCE
CLEANING
Perform the following cleaning procedures before use.
Pump: Wipe the pump with a clean, damp cloth and mild
detergent, and keep it away from dust. Air dry.
s NOTICE: Do not use phenolic products or corrosive
or powdered cleansers to clean the pump.
s NOTICE: Do not immerse or soak the pump.
Mattress: Wipe the mattress unit with a clean, damp cloth
and mild detergent. The mattress may also be cleaned
using a 10% solution of sodium hypochlorite (bleach)
diluted in water. Air dry all parts thoroughly before use.
s NOTICE: Do not use phenolic products to clean the
mattress.
s NOTICE: After cleaning, air dry the mattress without
direct exposure to sunlight.
GENERAL MAINTENANCE
1. Check main power cord and plug for abrasion or
excessive wear.
2. Check mattress cover for signs of wear or damage.
Ensure mattress cover and tubes are connected
correctly.
3. Check airflow from the air hose connector. The airflow
should alternate between each connector every half-
cycle time if it’s in alternating mode.
4. Check the air hoses for any kink or break. For
replacement, please contact your GF equipment
provider.
16
750000-INS-LAB-RevC11
Download from Www.Somanuals.com. All Manuals Search And Download.
Fuse replacement
tool needed: small screwdriver
1. If you suspect a blown fuse,
disconnect the plug from the
wall outlet immediately.
2. Use a small screwdriver to
remove the cover of the fuse
holder (fuse locations are shown
in picture at right).
3. Insert a new fuse of the correct
rating (T1A/250V, VDE approved).
4. Replace the fuse holder cover.
Air filter replacement
1. Remove the air filter plate
located at rear of pump (filter
location shown in picture at
right).
2. Remove the filter. The filter is
reusable, if not torn, and can
be washed gently with a mild
detergent and water. Dry the
filter before use.
3. Replace the air filter regularly if it is gray, torn, or the
environment is dirty.
STORAGE
1. Lay the mattress out flat and upside down.
2. Roll from the head end toward the foot end.
3. Stretch the foot-end strap around the rolled mattress to
prevent unrolling.
4. Store mattress and pump in a dust-free environment
with no exposure to direct sunlight.
s NOTICE: Do not fold, crease or stack the mattress.
750000-INS-LAB-RevC11
17
Download from Www.Somanuals.com. All Manuals Search And Download.
SPECIFICATIONS
Pump
Specification
Power Supply
120V System: AC 120V 60 Hz
Info: see rating label on product
Fuse Rating
Cycle time
Air input
T1A, 250V
8 minutes
>4 LPM
Dimensions (L x W x H)
Weight
9.8" x 5.3" x 3.7" (25 x 13.5 x 9.5 cm)
4 lb (1.8 kg)
Environment Temperature Operation: 50°F to 104°F (10°C to 40°C)
Storage: 5°F to 122°F ( -15°C to 50°C)
Shipping: 5°F to 158°F ( -15°C to 70°C)
Humidity
Operation: 10% to 90% non-condensing
Storage: 10% to 90% non-condensing
Shipping: 10% to 90% non-condensing
Classification
Class II, Type BF, IPX0
Applied Part: Air Mattress
Not suitable for use in the presence of a
flammable anesthetic mixture (No AP or
APG protection)
Mattress
Specification
8" mattress
Model
Dimensions (L x W x H)
Weight
78.7" x 35.4" x 8" (200 x 90 x 20.3 cm)
22.1 lb (10.0 kg)
20 x 8" bladders
Nylon/PU
Arrangement
Material
Cover
Cells
Base
Nylon/PU
PVC Tarpaulin
18
750000-INS-LAB-RevC11
Download from Www.Somanuals.com. All Manuals Search And Download.
WARRANTY
GF Health Products, Inc. warrants the Lumex AltaDyne
750000 Alternating Pressure / Low Air Loss Mattress
System for a period of twelve months for defects in
workmanship and materials. If a product is deemed to be
under warranty, GF Health Products, Inc. shall provide, at
its option, (1) replacement of any defective part or product
or (2) a credit of the original selling price made to GF
Health Products, Inc.’s initial customer. The warranty does
not include any labor charges incurred in replacement
part(s) installation or any associated freight or shipping
charges to GF Health Products, Inc.
The warranties contained herein contain all the
representations and warranties with respect to the
subject matter of this document, and supersede all
prior negotiations, agreements and understandings with
respect thereto. The recipient of this document hereby
acknowledges and represents that it has not relied on any
representation, assertion, guarantee, warranty, collateral
contract or other assurance, except those set out in this
document.
750000-INS-LAB-RevC11
19
Download from Www.Somanuals.com. All Manuals Search And Download.
USA Corporate Headquarters:
GF Health Products, Inc.
2935 Northeast Parkway
Atlanta, Georgia 30360
telephone: 770-368-4700
fax: 770-368-2386
© 2011, GF Health Products, Inc.