Graham Field Camping Equipment 750000 User Manual

750000  
ALTERNATING PRESSURE /  
LOW AIR LOSS MATTRESS SYSTEM  
USER MANUAL  
Important: Do not operate the Mattress System  
without first reading and understanding this manual!  
Save this manual for future use.  
Info: The most current version of this manual can be  
750000-INS-LAB-RevC11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
INTRODUCTION  
Use this manual for set-up and operation of the Lumex  
AltaDyne 750000 Alternating Pressure / Low Air Loss  
Mattress System. Read all instructions before using the  
mattress system. Save this manual for future reference.  
INTENDED USE OF THIS DEVICE  
The intended use of the Lumex AltaDyne 750000  
Alternating Pressure / Low Air Loss Mattress System is:  
• The treatment of pressure ulcers while optimizing user  
comfort.  
• Pain management as prescribed by a physician.  
• As described above, in either a homecare or long term  
care setting.  
Contraindication  
WARNING: DO NOT use this product in the presence  
of flammable anesthesia. There is a possible fire  
hazard when this product is used with certain oxygen  
delivery equipment.  
1
Use nasal cannula, face mask, or /2 length oxygen  
tent to deliver oxygen in the presence of this product.  
DO NOT use a full length oxygen tent that extends  
past the top surface of the mattress with this product.  
750000-INS-LAB-RevC11  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
IMPORTANT SAFETY PRECAUTIONS  
The safety statements presented in this chapter refer  
to the basic safety information that should be observed  
by those using the Alternating Pressure / Low Air Loss  
Mattress System. There are additional safety statements in  
other chapters or sections, which may be the same as or  
similar to the following, or specific to the operations.  
DANGER: Indicates an imminent hazard situation  
that, if not avoided, will result in death or serious  
injury.  
WARNING: Indicates a potential hazard situation or  
unsafe practice that, if not avoided, could result in  
death or serious injury.  
CAUTION: Indicates a potential hazard or unsafe  
practice that, if not avoided, could result in minor or  
moderate personal injury.  
s NOTICE: Indicates a potential hazard or unsafe  
practice that, if not avoided, could result in product  
or property damage.  
Info: Provides application recommendations or other useful  
information to ensure that you get the most from your  
product.  
DANGER  
To reduce the risk of electrocution:  
DANGER: Always unplug this product immediately  
after using.  
DANGER: Do not use this product while bathing.  
DANGER: Do not place or store this product where it  
can fall or be pulled into a tub or sink.  
DANGER: Do not place this product in or drop into  
water or other liquid.  
DANGER: Do not reach for a product that has fallen  
into water. Unplug it immediately.  
4
750000-INS-LAB-RevC11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARNING  
To reduce the risk of burns, electrocution, fire, or  
personal injury:  
WARNING: Do not leave this product unattended  
when plugged in.  
WARNING: Always use close supervision when this  
product is used by, on, or near children or those who  
require close supervision.  
WARNING: Use this product only as intended and  
described in this manual. Do not use attachments or  
accessories not recommended by Graham-Field.  
WARNING: Never operate this product if:  
a) It has a damaged cord or plug.  
b) It is not working properly.  
c) It has been dropped or damaged.  
d) It has been dropped into water.  
Return the product to your Graham-Field equipment  
provider.  
WARNING: Keep the cord away from heated surfaces.  
WARNING: Never drop or insert any object into any  
opening or hose.  
WARNING: Never block the air openings of this  
product or place it on a soft surface, such as a bed or  
couch, where the openings may be blocked. Keep the  
air opening free of lint, hair, and other similar particles.  
WARNING: Connect this product only to a properly  
grounded power outlet.  
WARNING: Maximum patient weight capacity for this  
product is 350 lb (159 kg).  
750000-INS-LAB-RevC11  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARNING: Patient entrapment with bed side rails  
may cause injury or death. The bed frame and its  
components, including the mattress, bed side rails,  
head and foot board, bedding, and any accessories  
added to the bed, can all affect the risk of entrapment.  
Thorough patient assessment and monitoring are  
necessary to reduce the risk of entrapment, including  
establishing whether the use of a bed rail is in the  
best interest of the patient. Read and understand the  
User Manual before using this equipment. GF Health  
Products, Inc. product manuals are available online at  
learn more about the risks of entrapment. It is the  
responsibility of the facility and provider to be in  
compliance with these guidelines. Refer to user  
manuals for beds and rails for additional product  
safety information.  
After any adjustment, repair or service, and before  
use, ensure all attaching hardware is securely  
tightened. Bed rails with dimensions different  
than the original equipment specified by the bed  
manufacturer may not be interchangeable and may  
result in entrapment or other injury.  
NOTICE  
s NOTICE: The pump can be used only with the  
accompanying mattress. Do not use it for any other  
purpose.  
6
750000-INS-LAB-RevC11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
EQUIPMENT SYMBOLS  
Attention! Consult accompanying documents (this manual)  
“BF”symbol: Indicates that this product is in accordance  
with the degree of protection against electric shock for type  
BF equipment  
Functional earth (for UL only)  
Class II (for UL only)  
Disposal of Electrical & Electronic Equipment (WEEE):  
Do not treat this product as household waste. For more  
detailed information with regard to returning and recycling  
this product, please consult your local city office, household  
waste disposal service, or Graham-Field equipment provider.  
For U.S. and Canada only  
Medical Equipment: Air Pump  
E228589  
53DG  
with respect to electrical shock, fire and  
mechanical hazards only in accordance with  
UL60601-1 and CAN/CSA C22.2 No. 601.1  
750000-INS-LAB-RevC11  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
PRODUCT DESCRIPTION  
PUMP AND MATTRESS SYSTEM  
foot end  
symbol  
CPR valves  
mattress  
pump  
mattress-pump  
connectors  
pump and mattress system  
Pump: front  
quick connector  
8
750000-INS-LAB-RevC11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Pump: rear  
Pump: top panel  
static function switch  
main power switch  
low pressure indicator  
normal pressure indicator  
pressure adjustment knob  
Pressure adjust knob: The pressure adjust knob controls  
the air pressure output. Turn it clockwise to increase  
output pressure (firmer); turn it counter-clockwise to  
decrease air output pressure (softer). Higher pressure  
output will better support the heavier-weight user. The  
setting is measured in kilograms (kg). Please consult your  
physician for an appropriate setting.  
Normal pressure indicator: The green normal pressure LED  
illuminates when the pressure inside the air mattress has  
reached the selected pressure setting.  
Low pressure indicator: The yellow low pressure LED  
illuminates when the pressure inside the air mattress is  
below the selected setting.  
750000-INS-LAB-RevC11  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Static function switch: Press the static function switch to  
suspend alternating function, if needed; the pressure inside  
the air tubes will then be adjusted to identical pressure.  
Press the static function switch again to change back  
to alternating mode. The static function switch toggles  
between static and alternating modes.  
Main power switch: The main power switch toggles  
between ON and OFF. When power is ON, either the normal  
or low pressure indicator illuminates.  
10  
750000-INS-LAB-RevC11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
INSTALLATION  
UNPACKING  
1. Before unpacking, check for obvious damage to the  
carton or its contents. If damage is evident, please  
notify the carrier and your Graham-Field equipment  
provider.  
2. Remove all loose packing from the carton.  
3. Carefully remove all components from the carton.  
4. Inspect all components. If damage is evident, please  
notify the carrier and your Graham-Field equipment  
provider.  
750000-INS-LAB-RevC11  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PUMP AND MATTRESS INSTALLATION  
1. Place the mattress on top  
of the bed frame; ensure  
that the foot end symbol  
faces up at foot end as  
shown at right.  
2. Fold open the wire  
hangers on the back of  
the pump as shown at  
right.  
3. Either hang the pump  
on the bed's foot end  
as shown at right, or  
place the pump on  
a flat surface easily  
accessible to the user,  
caregiver, and/or  
doctor.  
s NOTICE: Place the device in a position where the  
user, caregiver, and/or doctor can access it easily.  
4. Connect the mattress-  
pump air hose quick  
connector; ensure that  
the connector has clicked  
securely into place before  
continuing.  
s NOTICE: Ensure that the air hoses are not kinked or  
tucked under the mattress.  
12  
750000-INS-LAB-RevC11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARNING: Locate all cords so that they will not be  
stepped on, tripped over, or otherwise subjected to  
damage or stress.  
WARNING: Do not use a household extension cord if  
the electrical cord does not reach the power outlet.  
Use of an improper extension cord could result in  
fire and electric shock. If an extension cord must  
be used, use a three conductor cord with ground,  
properly wired, in good electrical condition, and keep  
it as short as possible.  
WARNING: Ensure that the local power voltage is  
appropriate for the pump unit.  
5. Plug the power cord into a properly grounded electrical  
outlet.  
750000-INS-LAB-RevC11  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OPERATION  
1. Turn the power switch  
to the ON position.  
2. Adjust the pressure  
adjust knob to the  
maximum position.  
Info: Set the pressure adjust knob to the maximum position  
for the quickest inflation at set-up; the mattress can then be  
adjusted to the desired pressure.  
3. The pump will begin  
delivering air into the  
mattress. While the  
mattress is filling, the  
low pressure indicator (yellow LED) will illuminate.  
4. When the selected  
pressure level is  
reached, the low  
pressure indicator will  
de-illuminate, and the normal pressure indicator (green  
LED) will automatically illuminate.  
5. Replace all sheets, blankets, and pillows on the bed.  
PRESSURE ADJUSTMENT  
Generally, a lighter patient will need a lower (softer)  
setting while a heavier patient will need a higher (firmer)  
setting, but pressure adjustment must ultimately be based  
on the patient's weight distribution. A short, stocky person  
may be the same weight as a tall, thin person, but their  
settings will not be the same. The mattress should be firm  
enough to support the patient, yet soft enough so the  
patient will sink down to eliminate pressure points. Test  
the pressure by placing four fingers below the mattress  
14  
750000-INS-LAB-RevC11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
under the patient's buttock area; your fingers should just  
fit. If the fit is too loose, decrease the setting. If your fingers  
do not fit well, increase the setting two notches firmer. Wait  
ten minutes and then check again. You should only have  
to adjust this setting one time per patient. Please consult  
your physician for an appropriate setting.  
QUICK DISCONNECT  
Pull the power plug from the wall connector to disconnect  
the device quickly.  
quick connector  
TRANSPORT CAP  
In case of power failure or transport:  
Disconnect the quick connector and  
cap the mattress with the transport  
cap to slow deflation.  
CPR FUNCTION  
transport cap  
If CPR must be performed on the  
patient while the mattress is in use,  
to immediately deflate the mattress,  
either:  
1. Pull the CPR valves from the  
mattress in the location of the  
arrow shown at right, or  
2. Disconnect the quick connector  
from the pump in the location of  
the arrow shown at right.  
750000-INS-LAB-RevC11  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MAINTENANCE  
CLEANING  
Perform the following cleaning procedures before use.  
Pump: Wipe the pump with a clean, damp cloth and mild  
detergent, and keep it away from dust. Air dry.  
s NOTICE: Do not use phenolic products or corrosive  
or powdered cleansers to clean the pump.  
s NOTICE: Do not immerse or soak the pump.  
Mattress: Wipe the mattress unit with a clean, damp cloth  
and mild detergent. The mattress may also be cleaned  
using a 10% solution of sodium hypochlorite (bleach)  
diluted in water. Air dry all parts thoroughly before use.  
s NOTICE: Do not use phenolic products to clean the  
mattress.  
s NOTICE: After cleaning, air dry the mattress without  
direct exposure to sunlight.  
GENERAL MAINTENANCE  
1. Check main power cord and plug for abrasion or  
excessive wear.  
2. Check mattress cover for signs of wear or damage.  
Ensure mattress cover and tubes are connected  
correctly.  
3. Check airflow from the air hose connector. The airflow  
should alternate between each connector every half-  
cycle time if it’s in alternating mode.  
4. Check the air hoses for any kink or break. For  
replacement, please contact your GF equipment  
provider.  
16  
750000-INS-LAB-RevC11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fuse replacement  
tool needed: small screwdriver  
1. If you suspect a blown fuse,  
disconnect the plug from the  
wall outlet immediately.  
2. Use a small screwdriver to  
remove the cover of the fuse  
holder (fuse locations are shown  
in picture at right).  
3. Insert a new fuse of the correct  
rating (T1A/250V, VDE approved).  
4. Replace the fuse holder cover.  
Air filter replacement  
1. Remove the air filter plate  
located at rear of pump (filter  
location shown in picture at  
right).  
2. Remove the filter. The filter is  
reusable, if not torn, and can  
be washed gently with a mild  
detergent and water. Dry the  
filter before use.  
3. Replace the air filter regularly if it is gray, torn, or the  
environment is dirty.  
STORAGE  
1. Lay the mattress out flat and upside down.  
2. Roll from the head end toward the foot end.  
3. Stretch the foot-end strap around the rolled mattress to  
prevent unrolling.  
4. Store mattress and pump in a dust-free environment  
with no exposure to direct sunlight.  
s NOTICE: Do not fold, crease or stack the mattress.  
750000-INS-LAB-RevC11  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SPECIFICATIONS  
Pump  
Specification  
Power Supply  
120V System: AC 120V 60 Hz  
Info: see rating label on product  
Fuse Rating  
Cycle time  
Air input  
T1A, 250V  
8 minutes  
>4 LPM  
Dimensions (L x W x H)  
Weight  
9.8" x 5.3" x 3.7" (25 x 13.5 x 9.5 cm)  
4 lb (1.8 kg)  
Environment Temperature Operation: 50°F to 104°F (10°C to 40°C)  
Storage: 5°F to 122°F ( -15°C to 50°C)  
Shipping: 5°F to 158°F ( -15°C to 70°C)  
Humidity  
Operation: 10% to 90% non-condensing  
Storage: 10% to 90% non-condensing  
Shipping: 10% to 90% non-condensing  
Classification  
Class II, Type BF, IPX0  
Applied Part: Air Mattress  
Not suitable for use in the presence of a  
flammable anesthetic mixture (No AP or  
APG protection)  
Mattress  
Specification  
8" mattress  
Model  
Dimensions (L x W x H)  
Weight  
78.7" x 35.4" x 8" (200 x 90 x 20.3 cm)  
22.1 lb (10.0 kg)  
20 x 8" bladders  
Nylon/PU  
Arrangement  
Material  
Cover  
Cells  
Base  
Nylon/PU  
PVC Tarpaulin  
18  
750000-INS-LAB-RevC11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARRANTY  
GF Health Products, Inc. warrants the Lumex AltaDyne  
750000 Alternating Pressure / Low Air Loss Mattress  
System for a period of twelve months for defects in  
workmanship and materials. If a product is deemed to be  
under warranty, GF Health Products, Inc. shall provide, at  
its option, (1) replacement of any defective part or product  
or (2) a credit of the original selling price made to GF  
Health Products, Inc.’s initial customer. The warranty does  
not include any labor charges incurred in replacement  
part(s) installation or any associated freight or shipping  
charges to GF Health Products, Inc.  
The warranties contained herein contain all the  
representations and warranties with respect to the  
subject matter of this document, and supersede all  
prior negotiations, agreements and understandings with  
respect thereto. The recipient of this document hereby  
acknowledges and represents that it has not relied on any  
representation, assertion, guarantee, warranty, collateral  
contract or other assurance, except those set out in this  
document.  
750000-INS-LAB-RevC11  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
USA Corporate Headquarters:  
GF Health Products, Inc.  
2935 Northeast Parkway  
Atlanta, Georgia 30360  
telephone: 770-368-4700  
fax: 770-368-2386  
© 2011, GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Graco Baby Jumper 1762254 User Manual
Grizzly Drill C1393 96 User Manual
Hotpoint Dishwasher BFV68 User Manual
Hotpoint Microwave Oven BU62 BU65 User Manual
HP Hewlett Packard Electronic Keyboard 3335A User Manual
HP Hewlett Packard Network Router 1740 SHDSL User Manual
HP Hewlett Packard Printer ws4500 User Manual
HP Hewlett Packard Server 8200zl 5400zl 3500yl Series User Manual
Hypertec Carrying Case N14568PHY User Manual
Iiyama Car Video System PLL400W PLL320W User Manual