Graham Field Blood Pressure Monitor 707A INS LAB RevD11 User Manual |
Item 707A,
707AX, 707AC
Digital Blood Pressure
Monitor
with Manual Inflation &
Automatic Deflation
User Manual
Read this manual before operating the Labtron Digital Blood
Pressure Monitor. Save this manual for future reference.
707A-INS-LAB-RevD11
Download from Www.Somanuals.com. All Manuals Search And Download.
SAFETY GUIDELINES - PLEASE READ BEFORE USE
Important! Read and understand these instructions before using the
Labtron Digital Blood Pressure Monitor. If you do not understand
any part of these instructions, contact your medical professional or
Graham-Field dealer for direction in the use of this product.
WARNING: If components are damaged or missing, contact your
Graham-Field dealer immediately. DO NOT use substitute parts.
WARNING: GF Health Products, Inc. assumes no responsibility for any
damage or injury caused by improper installation or use of this product.
INTENDED USE
The Labtron Digital Blood Pressure Monitor is intended to measure
arterial blood pressure. This device is intended for use by adults,
and not for use by children except under the supervision of an
adult. This device is not intended to replace regular medical
examinations. Review your procedure for using this monitor with
your physician, who is the only person qualified to interpret blood
pressure monitor results.
Labtron Digital Blood Pressure Monitor
3
Download from Www.Somanuals.com. All Manuals Search And Download.
FEATURES
bulb
Main features of
the Digital Blood
Pressure Monitor,
referred to in this
manual, are shown
in the picture at
right.
cuff
exhaust valve
button
LCD
cuff
hose
ON/OFF
button
Labtron Digital Blood Pressure Monitor
4
Download from Www.Somanuals.com. All Manuals Search And Download.
BATTERY INSTALLATION / REPLACEMENT
1. Slide the battery cover off in the direction of
the arrow.
N E P O
2. Install or replace the four AAA batteries,
oriented as shown by battery compartment
diagram, into the battery compartment.
3. Replace the battery cover. Ensure that the
battery cover is securely fastened.
Info: Ensure that the batteries are correctly
installed. Incorrect installation will prevent the
device from operating.
BLOOD PRESSURE
Blood Pressure is a measure of the blood’s pressure in the
circulatory system, which changes constantly during the course
of the cardiac cycle. Blood pressure readings report two values.
The higher reading (systolic pressure) shows the highest pressure
Labtron Digital Blood Pressure Monitor
5
Download from Www.Somanuals.com. All Manuals Search And Download.
in the arteries occurring when the heart contracts. The lower
reading (diastolic pressure) shows the lowest pressure in the
arteries, which occurs right before the heart contracts. Blood
pressure readings are written with the highest value first, then the
lowest value. Readings of 120/80 are considered to be normal, with
high blood pressure being defined as a systolic pressure which is
140mmHg or more at rest and a diastolic pressure which is 90mmHg
or more at rest. Only a patient’s physician is qualified to determine
whether the readings obtained are normal for that person.
Preparing to measure blood pressure
1. Blood pressure fluctuates continuously throughout the day. To
consistently track your blood pressure, try to take your blood
pressure at the same time each day.
2. Relax and remain still for 5 to 10 minutes before a blood pressure
measurement, and between blood pressure measurements.
3. Refrain from eating, smoking, and drinking, especially alcoholic
Labtron Digital Blood Pressure Monitor
6
Download from Www.Somanuals.com. All Manuals Search And Download.
beverages, before a measurement.
4. Remove any clothing on the upper arm to place the
cuff directly on the skin. Constriction of the upper
arm by clothing or a rolled-up sleeve, as shown at
right, may cause an inaccurate reading; remove any
constricting garments.
excessive
constriction
5. Sit down in a chair with back support
and place both feet on the floor.
6. Extend the arm from which the blood
pressure will be taken to the front or
side. Rest your hand, palm-up, on a
table as shown at right.
Info: Place the cuff at the same level as your heart to obtain accurate
measurements. Your heart is located slightly to the right of your left
armpit.
Labtron Digital Blood Pressure Monitor
7
Download from Www.Somanuals.com. All Manuals Search And Download.
7. Attach the cuff: Hold the area of the cuff where
the hose exits the cuff firmly against the inside
of the bare upper arm directly over the brachial
artery, and the bottom of the cuff approximately
one inch (2-3cm) above the elbow, as shown in the
third picture at right. Never place the cuff over
clothing. Pull the opposite end of the cuff snugly
around the arm and secure with the hook and loop
fastener. The cuff should be snug, but not too tight.
If one or two fingers can fit between the cuff and
the arm, the cuff is properly secured. If you are
unable to fit the cuff as described above, a smaller
or larger cuff should be used.
1 inch
Labtron Digital Blood Pressure Monitor
8
Download from Www.Somanuals.com. All Manuals Search And Download.
Measuring blood pressure
Symbol key: LCD symbols you may see during measurement
Deflating
Inflating
Measurement error
Measuring
Low battery
1. Ensure that cuff is properly placed and that you are positioned as
described in previous paragraphs.
2. Press the ON/OFF button. All LCD symbols will appear
momentarily. If a downward arrow symbol displays, press
the exhaust valve button (see earlier labeled picture) located
between the inflation bulb and the tube to release the residual
pressure from the cuff.
Labtron Digital Blood Pressure Monitor
9
Download from Www.Somanuals.com. All Manuals Search And Download.
Info: The low battery symbol will also appear at this time while the
monitor checks the power level. This does not indicate that the batteries
are low unless the low battery symbol remains on continuously.
3. After "0" displays, pressurize the cuff by squeezing the bulb until
the pressure value displayed is 50mmHg above your usual systolic
pressure. If you do not know your usual systolic pressure, squeeze
the bulb just until 200mmHg displays.
4. The cuff will automatically deflate after you stop pressurizing the
cuff. If an upward arrow symbol displays, you must inflate the
cuff to a higher pressure than you did the first time. If an upward
arrow does not display, the monitor is taking your measurement.
Remain calm and still to avoid causing any measurement error.
Diastolic
Systolic
5. The systolic pressure, diastolic pressure,
and pulse will display in the format shown
at right. The device will completely deflate
when measurement is finished. To accelerate
deflation, press the exhaust valve button.
Pulse rate
Labtron Digital Blood Pressure Monitor
10
Download from Www.Somanuals.com. All Manuals Search And Download.
Info: To stop measurement in progress for whatever reason, press the
exhaust valve button to release the pressure and turn off the device.
Info: If during measurement the power drops to an insufficient level,
the device will end the measurement and display the low battery
symbol. Replace the batteries and repeat the measurement.
Info: If the device cannot detect your pulse, it will end the
measurement. Wait several minutes, ensure that you and the cuff are
positioned properly, and try again.
6. The device will automatically shut off three minutes after
completing a reading, or you may turn it off manually by pressing
the ON/OFF button.
Labtron Digital Blood Pressure Monitor
11
Download from Www.Somanuals.com. All Manuals Search And Download.
7. If performing measurement for the first time, repeat the
measurement two or more times to ensure that you and cuff are
positioned correctly and that results are accurate and consistent.
Record your results; there is a log provided, with examples, at
the end of this manual for this purpose. Always check with your
physician to ensure that readings are performed correctly. If you
have concerns or questions about your results, always contact
your physician. Only your physician is qualified to analyze
blood pressure.
Labtron Digital Blood Pressure Monitor
12
Download from Www.Somanuals.com. All Manuals Search And Download.
MAINTENANCE
s NOTICE: To prevent product damage, please follow these
recommended care and maintenance instructions:
Do not drop or pull excessively on monitor components or subject them to strong impact.
Do not use monitor close to any device with a strong electrical field such as television,
microwave oven, X-Ray equipment, etc.
Never inflate cuff above 300mmHg. Never inflate cuff unless it is positioned on an arm.
Do not expose the monitor components to direct sunlight, high temperature, or high humidity.
Do not put any part of monitor in contact with sharp objects which could pierce the material
and cause damage.
Do not dismantle or disassemble monitor components.
Cleaning
Wipe off all components with a clean, dry cloth. Do not use solvents or other
petroleum-based cleaners. Do not press cuff with a hot iron.
Storage
Always deflate the cuff completely before storage.
Remove the batteries when the device is to be stored for extended periods of
time. Info: Removing the batteries will erase all readings in memory.
Labtron Digital Blood Pressure Monitor
13
Download from Www.Somanuals.com. All Manuals Search And Download.
TROUBLESHOOTING
Error indicators
Indication Cuff pressure reaches 330mmHg
Cause Twisted bladder inside cuff
Correction Ensure that bladder lies flat within cuff
Indication "Err" displays, cuff pressure releases, then "0 pressure" displays
Cause
Excessive arm movement or talking during measurement
Correction Turn monitor OFF then ON and measure again while remaining still
Indication Could not obtain pulse rate
Cause
Cuff was too loosely applied, extremely weak pulse, or arrhythmia
Correction Reapply cuff and measure again; if problem persists after cuff is
correctly positioned, seek medical attention immediately
Batteries are exhausted. Replace with four new AAA alkaline batteries
Labtron Digital Blood Pressure Monitor
14
Download from Www.Somanuals.com. All Manuals Search And Download.
General troubleshooting
Problem
Recommended Action
Nothing appears
in LCD
Ensure that batteries are correctly oriented in battery holder (polarity
matches battery case placement indicators)
No measurement Ensure that cuff is positioned correctly
occurs
Go over measurement procedure; measure again
Blood pressure
measurements are
too high or
Ensure that user and cuff are positioned as instructed
Rest your arm on a table so that the cuff is level with your heart during
measurement
Remain seated and still during entire measurement period
Refrain from hand and body movements during measurement
too low
Blood pressure
measurements
vary
Your blood pressure can fluctuate considerably throughout the day. All
of the following factors can influence your blood pressure:
• Emotional state
• Daily activities / exercise
• Smoking
• Drinking alcoholic beverages
• Eating
• Taking certain medications
Labtron Digital Blood Pressure Monitor
15
Download from Www.Somanuals.com. All Manuals Search And Download.
SPECIFICATIONS *
Digital Blood Pressure 707A, adult-size cuff, fits 7.5 in. - 12.2 in. (19-31cm) arm
Monitor Model
707A X, large adult-size cuff, fits 11.8 in. - 17.7 in. (30-45cm) arm
707A C, small adult-size cuff, fits 5.1 in. - 7.9 in. (13-29cm) arm
Measurement method
Display
Oscillometric
Digital LCD
Measurement range
Pressure 20-280mmHg, Pulse 40-180bpm
Measurement accuracy Pressure ±3mmHg, Pulse ±5%
Inflation system
Deflation system
Power
Manual inflation with air pumping bulb
Automatic deflation
1.5V (AAA) alkaline batteries X 4
Approximately 3 minutes after measurement
Auscultatory measurement
Automatic power-off
Reference method for
clinical trials
Weight
Approximately .75 lb (340g) without batteries
Labtron Digital Blood Pressure Monitor
16
Download from Www.Somanuals.com. All Manuals Search And Download.
Specifications, continued *
Environmental
conditions
Operation
50°F ~ 104°F (10°C ~ 40°C), 10% ~ 85% RH
14°F ~ 140°F (-10°C ~ 60°C), 10% ~ 95% RH
Storage &
transport
Standard cited
EN 1060-1
Noninvasive blood pressure measuring equipment
general requirements
EN 1060-3
Noninvasive blood pressure measuring equipment
supplementary requirements for electromechanical
blood pressure measuring systems
EN 60601-1 Safety requirements for medical electrical equipment
EN 60601-
1-2
Electromagnetic compatibility and safety for medical
electrical equipment
EN 14971
Risk analysis for medical devices
* Specifications are subject to change without notice
Labtron Digital Blood Pressure Monitor
17
Download from Www.Somanuals.com. All Manuals Search And Download.
WARRANTY
GF Health Products, Inc. ("Graham-Field") warrants the Labtron
Digital Blood Pressure Monitor Model 707A/707A X/707A C for a
period of one year for defects in workmanship and materials. If a
product is deemed to be under warranty, GF Health Products, Inc.
shall provide, at its option, (1) replacement of any defective part or
product or (2) a credit of the original selling price made to GF Health
Products, Inc.’s initial customer. The warranty does not include any
labor charges incurred in replacement part(s) installation or any
associated freight or shipping charges to GF Health Products, Inc.
The warranties contained herein contain all the representations
and warranties with respect to the subject matter of this
document, and supersede all prior negotiations, agreements and
understandings with respect thereto. The recipient of this document
hereby acknowledges and represents that it has not relied on any
representation, assertion, guarantee, warranty, collateral contract or
other assurance, except those set out in this document.
Labtron Digital Blood Pressure Monitor
18
Download from Www.Somanuals.com. All Manuals Search And Download.
BLOOD PRESSURE MEASUREMENT LOG
Date
Time
1/1
7:00
1/1
13:30
1/1
20:00
mmHg
240
220
200
180
160
140
120
100
80
134
90
128
84
123
76
69
60
Pulse
70
73
Body
Condition
Labtron Digital Blood Pressure Monitor
19
Download from Www.Somanuals.com. All Manuals Search And Download.
USA Corporate Headquarters:
GF Health Products, Inc.
2935 Northeast Parkway
Atlanta, Georgia 30360
telephone: 770-368-4700
fax: 770-368-2386
707A-INS-LAB-RevD11
© 2011, GF Health Products, Inc.
Download from Www.Somanuals.com. All Manuals Search And Download.
|