Graham Field Blood Pressure Monitor 707A INS LAB RevD11 User Manual

Item 707A,  
707AX, 707AC  
Digital Blood Pressure  
Monitor  
with Manual Inflation &  
Automatic Deflation  
User Manual  
Read this manual before operating the Labtron Digital Blood  
Pressure Monitor. Save this manual for future reference.  
707A-INS-LAB-RevD11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SAFETY GUIDELINES - PLEASE READ BEFORE USE  
Important! Read and understand these instructions before using the  
Labtron Digital Blood Pressure Monitor. If you do not understand  
any part of these instructions, contact your medical professional or  
Graham-Field dealer for direction in the use of this product.  
WARNING: If components are damaged or missing, contact your  
Graham-Field dealer immediately. DO NOT use substitute parts.  
WARNING: GF Health Products, Inc. assumes no responsibility for any  
damage or injury caused by improper installation or use of this product.  
INTENDED USE  
The Labtron Digital Blood Pressure Monitor is intended to measure  
arterial blood pressure. This device is intended for use by adults,  
and not for use by children except under the supervision of an  
adult. This device is not intended to replace regular medical  
examinations. Review your procedure for using this monitor with  
your physician, who is the only person qualified to interpret blood  
pressure monitor results.  
Labtron Digital Blood Pressure Monitor  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
FEATURES  
bulb  
Main features of  
the Digital Blood  
Pressure Monitor,  
referred to in this  
manual, are shown  
in the picture at  
right.  
cuff  
exhaust valve  
button  
LCD  
cuff  
hose  
ON/OFF  
button  
Digital Blood Pressure Monitor  
Labtron Digital Blood Pressure Monitor  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
BATTERY INSTALLATION / REPLACEMENT  
1. Slide the battery cover off in the direction of  
the arrow.  
N E P O  
2. Install or replace the four AAA batteries,  
oriented as shown by battery compartment  
diagram, into the battery compartment.  
3. Replace the battery cover. Ensure that the  
battery cover is securely fastened.  
Info: Ensure that the batteries are correctly  
installed. Incorrect installation will prevent the  
device from operating.  
BLOOD PRESSURE  
Blood Pressure is a measure of the blood’s pressure in the  
circulatory system, which changes constantly during the course  
of the cardiac cycle. Blood pressure readings report two values.  
The higher reading (systolic pressure) shows the highest pressure  
Labtron Digital Blood Pressure Monitor  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
in the arteries occurring when the heart contracts. The lower  
reading (diastolic pressure) shows the lowest pressure in the  
arteries, which occurs right before the heart contracts. Blood  
pressure readings are written with the highest value first, then the  
lowest value. Readings of 120/80 are considered to be normal, with  
high blood pressure being defined as a systolic pressure which is  
140mmHg or more at rest and a diastolic pressure which is 90mmHg  
or more at rest. Only a patient’s physician is qualified to determine  
whether the readings obtained are normal for that person.  
Preparing to measure blood pressure  
1. Blood pressure fluctuates continuously throughout the day. To  
consistently track your blood pressure, try to take your blood  
pressure at the same time each day.  
2. Relax and remain still for 5 to 10 minutes before a blood pressure  
measurement, and between blood pressure measurements.  
3. Refrain from eating, smoking, and drinking, especially alcoholic  
Labtron Digital Blood Pressure Monitor  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
beverages, before a measurement.  
4. Remove any clothing on the upper arm to place the  
cuff directly on the skin. Constriction of the upper  
arm by clothing or a rolled-up sleeve, as shown at  
right, may cause an inaccurate reading; remove any  
constricting garments.  
excessive  
constriction  
5. Sit down in a chair with back support  
and place both feet on the floor.  
6. Extend the arm from which the blood  
pressure will be taken to the front or  
side. Rest your hand, palm-up, on a  
table as shown at right.  
Info: Place the cuff at the same level as your heart to obtain accurate  
measurements. Your heart is located slightly to the right of your left  
armpit.  
Labtron Digital Blood Pressure Monitor  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Attach the cuff: Hold the area of the cuff where  
the hose exits the cuff firmly against the inside  
of the bare upper arm directly over the brachial  
artery, and the bottom of the cuff approximately  
one inch (2-3cm) above the elbow, as shown in the  
third picture at right. Never place the cuff over  
clothing. Pull the opposite end of the cuff snugly  
around the arm and secure with the hook and loop  
fastener. The cuff should be snug, but not too tight.  
If one or two fingers can fit between the cuff and  
the arm, the cuff is properly secured. If you are  
unable to fit the cuff as described above, a smaller  
or larger cuff should be used.  
1 inch  
Labtron Digital Blood Pressure Monitor  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Measuring blood pressure  
Symbol key: LCD symbols you may see during measurement  
Deflating  
Inflating  
Measurement error  
Measuring  
Low battery  
1. Ensure that cuff is properly placed and that you are positioned as  
described in previous paragraphs.  
2. Press the ON/OFF button. All LCD symbols will appear  
momentarily. If a downward arrow symbol displays, press  
the exhaust valve button (see earlier labeled picture) located  
between the inflation bulb and the tube to release the residual  
pressure from the cuff.  
Labtron Digital Blood Pressure Monitor  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Info: The low battery symbol will also appear at this time while the  
monitor checks the power level. This does not indicate that the batteries  
are low unless the low battery symbol remains on continuously.  
3. After "0" displays, pressurize the cuff by squeezing the bulb until  
the pressure value displayed is 50mmHg above your usual systolic  
pressure. If you do not know your usual systolic pressure, squeeze  
the bulb just until 200mmHg displays.  
4. The cuff will automatically deflate after you stop pressurizing the  
cuff. If an upward arrow symbol displays, you must inflate the  
cuff to a higher pressure than you did the first time. If an upward  
arrow does not display, the monitor is taking your measurement.  
Remain calm and still to avoid causing any measurement error.  
Diastolic  
Systolic  
5. The systolic pressure, diastolic pressure,  
and pulse will display in the format shown  
at right. The device will completely deflate  
when measurement is finished. To accelerate  
deflation, press the exhaust valve button.  
Pulse rate  
Labtron Digital Blood Pressure Monitor  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Info: To stop measurement in progress for whatever reason, press the  
exhaust valve button to release the pressure and turn off the device.  
Info: If during measurement the power drops to an insufficient level,  
the device will end the measurement and display the low battery  
symbol. Replace the batteries and repeat the measurement.  
Info: If the device cannot detect your pulse, it will end the  
measurement. Wait several minutes, ensure that you and the cuff are  
positioned properly, and try again.  
6. The device will automatically shut off three minutes after  
completing a reading, or you may turn it off manually by pressing  
the ON/OFF button.  
Labtron Digital Blood Pressure Monitor  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. If performing measurement for the first time, repeat the  
measurement two or more times to ensure that you and cuff are  
positioned correctly and that results are accurate and consistent.  
Record your results; there is a log provided, with examples, at  
the end of this manual for this purpose. Always check with your  
physician to ensure that readings are performed correctly. If you  
have concerns or questions about your results, always contact  
your physician. Only your physician is qualified to analyze  
blood pressure.  
Labtron Digital Blood Pressure Monitor  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
MAINTENANCE  
s NOTICE: To prevent product damage, please follow these  
recommended care and maintenance instructions:  
Do not drop or pull excessively on monitor components or subject them to strong impact.  
Do not use monitor close to any device with a strong electrical field such as television,  
microwave oven, X-Ray equipment, etc.  
Never inflate cuff above 300mmHg. Never inflate cuff unless it is positioned on an arm.  
Do not expose the monitor components to direct sunlight, high temperature, or high humidity.  
Do not put any part of monitor in contact with sharp objects which could pierce the material  
and cause damage.  
Do not dismantle or disassemble monitor components.  
Cleaning  
Wipe off all components with a clean, dry cloth. Do not use solvents or other  
petroleum-based cleaners. Do not press cuff with a hot iron.  
Storage  
Always deflate the cuff completely before storage.  
Remove the batteries when the device is to be stored for extended periods of  
time. Info: Removing the batteries will erase all readings in memory.  
Labtron Digital Blood Pressure Monitor  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
TROUBLESHOOTING  
Error indicators  
Indication Cuff pressure reaches 330mmHg  
Cause Twisted bladder inside cuff  
Correction Ensure that bladder lies flat within cuff  
Indication "Err" displays, cuff pressure releases, then "0 pressure" displays  
Cause  
Excessive arm movement or talking during measurement  
Correction Turn monitor OFF then ON and measure again while remaining still  
Indication Could not obtain pulse rate  
Cause  
Cuff was too loosely applied, extremely weak pulse, or arrhythmia  
Correction Reapply cuff and measure again; if problem persists after cuff is  
correctly positioned, seek medical attention immediately  
Batteries are exhausted. Replace with four new AAA alkaline batteries  
Labtron Digital Blood Pressure Monitor  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
General troubleshooting  
Problem  
Recommended Action  
Nothing appears  
in LCD  
Ensure that batteries are correctly oriented in battery holder (polarity  
matches battery case placement indicators)  
No measurement Ensure that cuff is positioned correctly  
occurs  
Go over measurement procedure; measure again  
Blood pressure  
measurements are  
too high or  
Ensure that user and cuff are positioned as instructed  
Rest your arm on a table so that the cuff is level with your heart during  
measurement  
Remain seated and still during entire measurement period  
Refrain from hand and body movements during measurement  
too low  
Blood pressure  
measurements  
vary  
Your blood pressure can fluctuate considerably throughout the day. All  
of the following factors can influence your blood pressure:  
• Emotional state  
• Daily activities / exercise  
• Smoking  
• Drinking alcoholic beverages  
• Eating  
• Taking certain medications  
Labtron Digital Blood Pressure Monitor  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SPECIFICATIONS *  
Digital Blood Pressure 707A, adult-size cuff, fits 7.5 in. - 12.2 in. (19-31cm) arm  
Monitor Model  
707A X, large adult-size cuff, fits 11.8 in. - 17.7 in. (30-45cm) arm  
707A C, small adult-size cuff, fits 5.1 in. - 7.9 in. (13-29cm) arm  
Measurement method  
Display  
Oscillometric  
Digital LCD  
Measurement range  
Pressure 20-280mmHg, Pulse 40-180bpm  
Measurement accuracy Pressure ±3mmHg, Pulse ±5%  
Inflation system  
Deflation system  
Power  
Manual inflation with air pumping bulb  
Automatic deflation  
1.5V (AAA) alkaline batteries X 4  
Approximately 3 minutes after measurement  
Auscultatory measurement  
Automatic power-off  
Reference method for  
clinical trials  
Weight  
Approximately .75 lb (340g) without batteries  
Labtron Digital Blood Pressure Monitor  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications, continued *  
Environmental  
conditions  
Operation  
50°F ~ 104°F (10°C ~ 40°C), 10% ~ 85% RH  
14°F ~ 140°F (-10°C ~ 60°C), 10% ~ 95% RH  
Storage &  
transport  
Standard cited  
EN 1060-1  
Noninvasive blood pressure measuring equipment  
general requirements  
EN 1060-3  
Noninvasive blood pressure measuring equipment  
supplementary requirements for electromechanical  
blood pressure measuring systems  
EN 60601-1 Safety requirements for medical electrical equipment  
EN 60601-  
1-2  
Electromagnetic compatibility and safety for medical  
electrical equipment  
EN 14971  
Risk analysis for medical devices  
* Specifications are subject to change without notice  
Labtron Digital Blood Pressure Monitor  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
WARRANTY  
GF Health Products, Inc. ("Graham-Field") warrants the Labtron  
Digital Blood Pressure Monitor Model 707A/707A X/707A C for a  
period of one year for defects in workmanship and materials. If a  
product is deemed to be under warranty, GF Health Products, Inc.  
shall provide, at its option, (1) replacement of any defective part or  
product or (2) a credit of the original selling price made to GF Health  
Products, Inc.’s initial customer. The warranty does not include any  
labor charges incurred in replacement part(s) installation or any  
associated freight or shipping charges to GF Health Products, Inc.  
The warranties contained herein contain all the representations  
and warranties with respect to the subject matter of this  
document, and supersede all prior negotiations, agreements and  
understandings with respect thereto. The recipient of this document  
hereby acknowledges and represents that it has not relied on any  
representation, assertion, guarantee, warranty, collateral contract or  
other assurance, except those set out in this document.  
Labtron Digital Blood Pressure Monitor  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
BLOOD PRESSURE MEASUREMENT LOG  
Date  
Time  
1/1  
7:00  
1/1  
13:30  
1/1  
20:00  
mmHg  
240  
220  
200  
180  
160  
140  
120  
100  
80  
134  
90  
128  
84  
123  
76  
69  
60  
Pulse  
70  
73  
Body  
Condition  
Labtron Digital Blood Pressure Monitor  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
USA Corporate Headquarters:  
GF Health Products, Inc.  
2935 Northeast Parkway  
Atlanta, Georgia 30360  
telephone: 770-368-4700  
fax: 770-368-2386  
707A-INS-LAB-RevD11  
© 2011, GF Health Products, Inc.  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Graco Car Seat PD261473A User Manual
Graco Car Seat PD262895A User Manual
Grizzly Sander T26417 T26418 User Manual
Harman Stove Company Stove 929BV User Manual
Havis Shields Automobile Accessories C TSM CV D User Manual
Honeywell Security Camera HZC 855NX User Manual
HP Hewlett Packard Calculator 48G User Manual
HP Hewlett Packard Laptop VC4815 User Manual
HP Hewlett Packard Server BL870c User Manual
Ikelite Digital Camera DSC V3 User Manual