F-801 INTERNAL REVERSE OSMOSIS UNIT  
Installation, Operation and Maintenance Manual  
TABLE OF CONTENTS  
Page  
1-1  
CHAPTER ONE:  
GENERAL INFORMATION  
Section  
1.1  
1.2  
1.3  
1.4  
1.5  
1.6  
1.7  
1.8  
1.9  
The Manual  
Safety Summary  
Applications  
Contraindications  
Environmental Considerations  
Theory of Operation  
Fluid Components  
Monitors and Controls  
Optional Equipment  
1-3  
1-3  
1-4  
1-4  
1-5  
1-6  
1-7  
1-9  
1-11  
1-12  
1.10  
Specifications  
1.10.1  
1.10.2  
1.10.3  
1.10.4  
1.10.5  
1.11  
1.12  
1.13  
1.14  
1.15  
Inlet Water Requirements  
Flow Rates  
Electrical Requirements  
Membrane Performance Characteristics  
Environmental Requirements  
Disposal  
Electromagnetic Interference  
Service Assistance  
Return Goods Authorization (RGA)  
Symbols and Abbreviations  
1-13  
1-13  
1-14  
1-15  
1-16  
CHAPTER TWO:  
INSTALLATION GUIDELINES AND INITIAL  
SYSTEM START UP  
2-1  
Section  
2.1  
2.2  
2.3  
Installation Considerations  
Installation Procedure  
Product Water Analysis Procedure  
2-3  
2-3  
2-5  
CHAPTER THREE:  
SYSTEM OPERATION  
3-1  
Section  
3.1  
3.2  
3.3  
Daily Start-up Procedure  
Bicarbonate Container Filling  
Shut Down Procedure  
3-3  
3-6  
3-6  
12552e  
A
27 February 2004  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CHAPTER FOUR:  
ROUTINE MAINTENANCE  
4-1  
Section  
4.1  
4.1.1  
Routine Maintenance of F-801 Internal RO  
System  
4-3  
4-3  
Daily / Per Treatment Maintenance  
4.1.2  
4.1.3  
4.1.4  
4.1.5  
Weekly Maintenance  
Monthly Maintenance  
Quarterly Maintenance  
Annual Maintenance  
4-5  
4-5  
4-6  
4-6  
4.2  
Specific Maintenance Procedures and Instructions 4-7  
4.2.1  
4.2.2  
4.2.3  
Motor Lubricant Procedure  
Leak Repair  
Pre-filter Replacement  
4-7  
4-7  
4-9  
4.3  
4.3.1  
F-801 Cleaning Indications  
Pre-Cleaning Preparation  
Pre-Cleaning Procedure  
Cleaner Recirculation and Dwell Time  
Post-Cleaning Rinse Procedure  
4-11  
4-12  
4-12  
4-13  
4-14  
4-16  
4-16  
4-17  
4-18  
4-20  
4-20  
4-21  
4-22  
4-23  
4-24  
4.3.1.1  
4.3.1.2  
4.3.1.3  
4.3.2  
4.3.2.1  
4.3.2.2  
4.3.2.3  
4.4  
BIOSAN Pre Cleaning Preparation  
BIOSAN Cleaning Procedure  
BIOSAN Recirculation and Dwell Time  
BIOSAN Post Cleaning Rinse Procedure  
F-801 Disinfection Indications  
Formaldehyde Disinfection Preparation  
Formaldehyde Pre-Disinfection Procedure  
Formaldehyde Disinfection and Dwell Time  
Formaldehyde Post-Disinfection Rinse Procedure  
“RENALIN®” Disinfection Procedure  
(Peracetic Acid and Peroxide-Based Compounds  
such as Renalin®, Minncare, & Peracidin)  
“RENALIN®” Pre-Disinfection Preparation  
“RENALIN®” Disinfection and Dwell Time  
“RENALIN®” Post-Disinfection Rinse Procedure  
Product Water Culture  
4.4.1  
4.4.1.1  
4.4.1.2  
4.4.1.3  
4.4.2  
4.4.2.1  
4.4.2.2  
4.4.2.3  
4.5  
4.5.1  
4.6  
4-25  
4-26  
4-28  
4-29  
4-30  
4-31  
4-31  
4-32  
Product Water Culture Procedure  
System Storage  
Formaldehyde Storage Procedure  
BIOSAN Storage Procedure  
4.6.1  
4.6.2  
12552e  
B
27 February 2004  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CHAPTER FIVE:  
TROUBLESHOOTING  
5-1  
Table of Content  
RO Fails to Run  
RO Runs Intermittently  
Motor Runs Hot  
Low Pump Pressure  
Pressure Gauge Vibrates Erratically  
Excess Product Pressure (too high)  
Product Flow Too Low  
Product Flow Too High  
Membrane Product Pressure Too Low  
Leaks  
Low Percent Rejection  
Rejection Display Fluctuates  
TDS Value High  
5-3  
5-5  
5-5  
5-6  
5-6  
5-7  
5-7  
5-7  
5-8  
5-9  
5-9  
5-9  
5-10  
5-10  
5-11  
5-12  
Section  
5.1  
5.2  
5.3  
5.4  
5.5  
5.6  
5.7  
5.8  
5.9  
5.10  
5.11  
5.12  
5.13  
5.14  
5.15  
RO Fails to Draw Disinfectant  
RO Pump Leaks  
CHAPTER SIX:  
DRAWINGS  
6-1  
7-1  
CHAPTER SEVEN:  
SPARE PARTS LIST  
12552e  
C
27 February 2004  
Download from Www.Somanuals.com. All Manuals Search And Download.  
This page intentionally left blank  
12552e  
D
27 February 2004  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
F-801 Internal RO System  
CHAPTER ONE:  
GENERAL INFORMATION  
12552e – 27 Feb 04  
1 - 1  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e – 27 Feb 04  
1 - 2  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
1.1  
The Manual  
This manual has been prepared to provide the operator with information and  
instructions regarding the installation, use, maintenance and troubleshooting of the GE  
Osmonics F-801 Internal Reverse Osmosis System.  
CAUTION:  
When used as a medical device, Federal law restricts this device to sale by or on  
the order of a physician. Per CFR 801.109 (b)(1).  
The manual has been written in narrative form supplemented with schematics and  
drawings for clarification. The operator can perform most procedures mentioned in  
this manual. Any exceptions will be clearly identified by a qualifying statement.  
1.2 Safety Summary  
Words in BOLD CAPITAL letters are used to identify labels on the device and key  
safety or qualifying statements. A list of all symbols and abbreviations is located at the  
end of this chapter. See Technote 120 for a glossary of relevant terms.  
This safety summary does not contain all of the safety statements in the manual. Other  
safety statements are included within the manual text and are enhanced and defined as  
follows:  
NOTE:  
Statements that provide further clarification.  
CAUTION:  
Statements identifying conditions or practices that could result in equipment or  
other property damage.  
WARNING:  
Statements identifying conditions or practices that could result in personal injury  
or loss of life.  
READ THIS MANUAL:  
Prior to operating or servicing this device, this manual must be read and understood.  
Keep this and other associated manuals for future reference and for new operators or  
qualified service personnel. A note sheet is provided at the end of each chapter for  
operators to make notations that may be valuable to other users.  
USE PROPER POWER CONNECTIONS:  
Use proper wiring and connection methods to satisfy hospital electrical codes.  
DO NOT REMOVE COVERS OR PANELS:  
To avoid electrical shock hazard, do not remove covers or panels when power is  
supplied to the device. Do not operate the device when covers or panels are removed.  
12552e – 27 Feb 04  
1 - 3  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
USE PROPER WATER AND DRAIN CONNECTIONS:  
Use proper water and drain connections to prevent contamination of the RO and danger  
to the patient. To ease confusion, the Quick Disconnect (QD) fittings have been color-  
coded. Both the male and female fittings are colored. The color coding is; Green =  
FEED water, Red = WASTE water, White = PERMEATE (Product) water.  
SHOCK HAZARD:  
Connect this device to a proper ground connection in accordance with the National  
Electrical Code. DO NOT under any circumstances remove the ground wire or ground  
prong from any power plug. DO NOT use an extension cord with this equipment.  
Keep the surface under the RO(s) free of liquids. Fix any leaks from the hemodialysis  
unit promptly.  
DEVICE LABELING:  
Do not under any circumstances; remove any Caution, Warning or any other  
descriptive labels from the devices until the conditions warranting the label are  
eliminated.  
DO NOT OPERATE IN A FLAMMABLE ATMOSPHERE:  
To avoid fire or explosion, do not operate this device in an explosive environment or  
near flammable anesthetics.  
DO NOT TRY TO CATCH THE EQUIPMENT IF IT STARTS TO FALL OVER:  
The overall assembly is weighty and can cause physical harm.  
1.3 Applications  
The F-801 Internal Reverse Osmosis (RO) System is designed to purify water by  
forcing water through a semi-permeable RO membrane. Water purified by reverse  
osmosis has had approximately 95% of the dissolved ions and approximately 99% of  
most other contaminants removed. The quality of the purified water, referred to as  
product water or permeate, depends on the quality of the feed water and normally is  
equal to or exceeds AAMI standards. Such water is suitable for hemodialysis  
applications.  
1.4 Contraindications  
WARNING:  
GE Osmonics systems are not designed, sold, or intended for use in producing  
water for injection  
Furthermore, GE Osmonics systems are not intended to be used outside of the device  
specifications and limitations, as outlined in this manual and other related materials.  
12552e – 27 Feb 04  
1 - 4  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
1.5 Environmental Considerations  
Prior to the installation of the F-801 RO System, it will be necessary to provide utilities  
and create an environment suitable for the trouble free operation of the RO system and  
its accessories.  
POWER:  
The F-801 operates on 115 VAC single-phase power. Histories of power failure,  
power surges, and low line voltages should be noted and reported to the  
manufacturer or their agent as they may create adverse conditions for the  
operation of equipment.  
WATER:  
Reverse osmosis systems require a reliable water source. The F-801 RO needs a  
minimum of 30-PSI pressure at a minimum flow of 1.6 gallons per minute. It is  
important that water is flowing at the design flow rate when testing water pressure  
rather than reading static pressure.  
DRAIN:  
A drain outlet is required by the RO system. The drain must have a minimum  
capacity of 1.6 gallons per RO per minute of continuous flow for as long as and  
as often as the RO is operating.  
TEMPERATURE:  
The F-801 RO is normally connected to a cold water supply. The performance  
specifications of the RO are based on 77°F (25°C) feed temperature. Each degree  
Fahrenheit the feed water temperature falls, the RO product flow decreases  
approximately 1.5%, and each degree Celsius drop, the product flow is reduced  
by 3% (See Technote 113 for the “Temperature Correction Factors” chart). In  
climates where the winter water temperature drops significantly below 77°F it  
may be necessary to temper (heat) the water.  
WARNING:  
The temperature of the tempered water should be monitored closely with a  
thermometer as high temperatures can damage the RO membranes and possibly  
harm the patient. Refer to the inlet water requirements (Section 1.10.1) for the  
maximum operating temperature for the F-801 RO.  
CAUTION:  
It is necessary to keep the RO from freezing to prevent serious damage to the RO  
unit.  
NOISE:  
The RO unit does not generate significant noise levels. Hard walls may reflect  
noise and will make the RO seem louder.  
12552e – 27 Feb 04  
1 - 5  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
WATER PURITY:  
Plumbing materials can significantly contribute to the contamination of the water.  
Metallic pipes (with the exception of 316 stainless steel) must be eliminated once  
the purification process has begun. Schedule 40 or 80 PVC pipes, polypropylene,  
PVDF, and other FDA recommended materials are suitable for most grades of  
purified water. Care must also be exercised in the choice of a thread sealant.  
Teflon tape is suitable for all threaded connections in this system.  
WARNING:  
Ordinary pipe dope must be avoided since it may leach objectionable and  
potentially dangerous impurities into the water.  
1.6  
Theory of Operation  
The F-801 Portable Reverse Osmosis system (RO) purifies a given feed water (by way  
of reverse osmosis) for use in hemodialysis applications. The F-801 RO possesses  
monitoring for pressure, product water quality and membrane performance (percent  
rejection). The unit includes pre-filtration to remove particulates greater than 5  
microns, which could clog orifices or damage components. The RO pump provides the  
pressure required to push water through the RO membrane. The fluid controls provide  
a means of managing flow rates and pressures. The following paragraphs detail the  
function of these fluid control components, which are used within the RO.  
12552e – 27 Feb 04  
1 - 6  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
1.7 Fluid Components  
Pre-filter:  
From the feed water, the fluid path flows through the softener (optional) on  
through the carbon tank and then into the pre-filter, located in a blue housing  
mounted on the side of the hemodialysis unit. It is a cartridge type filter with a  
nominal filtration rating of five microns.  
Pressure Adjust Regulator:  
Used to adjust the product pressure. The pressure is displayed on the product  
pressure gauge, on the front of the RO.  
High-Pressure Pump:  
Provides up to a 200 PSI boost of the inlet water to the Reverse Osmosis  
Membrane Assembly (ROMA). The RO pump pressure is displayed on a 0 to  
300 PSI gauge mounted outside the RO, below the product pressure gauge. The  
pressure is factory set at 150 PSI.  
Reverse Osmosis Membrane Assembly (ROMA):  
One is included in the F-801 RO. A ROMA is an assembly that includes the  
membrane housing, end caps, fittings, and a membrane. The thin film membrane  
will produce 600 (±15%) gallons per day of product water, depending on  
membrane condition and water temperature.  
Waste Flow:  
The Waste Flow is controlled by a length of 1/8" capillary tubing. The length of  
the tube is factory determined to correctly govern the wastewater flow to drain.  
CAUTION:  
This tube must not be shortened.  
Dialyze/Flush Valve:  
The Dialyze/Flush Valve is used to flush the membrane surface of rejected  
contaminants, and to accelerate the rinse out of cleaners or disinfectants from the  
RO. Some pump noise (chattering) is normal in the flush position. The RO is in  
"flush" from the full open (flush) to just before the fully closed (dialyze) position.  
CAUTION:  
Excessive “chattering” may be eliminated by slowly moving the valve from the  
full flush position towards “DIALYZE”.  
Product Recirculation Check Valve:  
Prevents over-pressuring of the product supply. When the product pressure  
exceeds the set point of the pressure regulator, the product recirculation check  
valve opens. Excess product water is then recirculated back to the inlet water  
flow.  
12552e – 27 Feb 04  
1 - 7  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
Feed Port:  
The Feed Port is a green colored quick-disconnection port for the introduction of  
feed water to the RO. The port has a check valve that closes when there is no  
hose connected to the fitting.  
Product Port:  
The Product Port is a white colored quick-disconnection port for the output of  
purified product water. The port has a check valve that closes when there is no  
hose connected to the fitting.  
Waste Port:  
The Waste Port is a red colored quick-disconnection port for the output of  
concentrate or waste to the drain. The port has a check valve that closes when  
there is no hose connected to the fitting.  
12552e – 27 Feb 04  
1 - 8  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
1.8 Monitors and Controls  
RO Power Switch:  
The RO Power Switch is mounted on the back of the remote monitor module.  
The switch is used to turn the F-801 RO unit on and off.  
Inlet Water Sensor:  
Monitors the quality of the inlet water. Inlet quality is compared to the outlet  
(product) water quality and is displayed as "percent rejection" on the front panel.  
Product Water Sensor:  
The Product Water Sensor is used in conjunction with the inlet water sensor by  
the percent-rejection circuitry to calculate the membrane rejection rate. The rate  
of rejection is displayed from 70% - 99% in a 5 LED array on the front panel.  
Total Dissolved Solids (TDS) Sensor:  
Samples the product water quality via the sensor probe, and the final quality of the  
product water is indicated on a digital display on the front panel as TDS in mg/L  
and ppm. The monitor is temperature compensated to adjust the display value to  
changing water temperature.  
Pressure Adjust Regulator:  
The Pressure Adjust Regulator inside the RO is used to set the maximum product  
pressure needed to supply a dialysis machine. Adjusting this regulator will also  
result in small adjustments in the pump pressure.  
Pre and Post filter Pressure Gauges:  
The Pre-filter pressure gauges are mounted on top of the pre-filter housing and  
display the feed water pressure from zero to 100 PSI. The two gauges are used to  
determine the feed water pressure and when compared against each other to  
determine fouling of the filter (delta pressure).  
Product Pressure Gauge:  
The Product Pressure Gauge reads the product pressure from 0 to 100 PSI and is  
located on the RO module right side panel. Product pressure will read zero (or  
close to) when the product line is not connected to a dialysis unit and the product  
hose is connected to the product fitting and the valve is open. The Fresenius  
dialysis machine requires a minimum inlet pressure of 20 PSI.  
Pump Pressure Gauge:  
The Pump Pressure Gauge reads the pump pressure from 0 to 300 PSI and is  
located on the RO module right side panel. It is used for monitoring during  
operation and when adjusting the pump pressure, and is located on the front of the  
RO.  
12552e – 27 Feb 04  
1 - 9  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
Dialyze/Flush Valve:  
The Dialyze/Flush Valve is a 1/4-turn ball valve. When in the dialyze mode, the  
valve is in the closed position. When the valve is turned at any point towards  
flush, the valve is open. The closer to the flush point, the more the valve is open.  
When in flush, the valve bypasses the waste flow tube.  
Percent Rejection Display:  
The Percent Rejection Display is a 5 LED array that indicates how the RO  
membranes are performing. The lights correspond to 99%, 95%, 90%, 85%, and  
>70% rejection. The 99% light, when illuminated, indicates a need to evaluate  
the RO system performance. RO units do not normally perform in excess of  
99%; verify this reading with an independent meter and take corrective action if  
indicated. Some RO membranes will achieve this level of performance under  
optimum conditions.  
Product Water Total Dissolved Solids (TDS) Display:  
The Product Water Total Dissolved Solids (TDS) digital display indicates product  
water purity in mg/L (ppm). It is temperature compensated to adjust the value of  
the display with changing water temperature for better accuracy.  
Less Than 0.5 mg/L Check System Lamp:  
The Less Than 0.5 mg/L Check System lamp illuminates when the TDS reading is  
0.5 mg/L or less. This indicates a need to evaluate the RO system performance.  
RO units do not normally have this low product TDS. Verify this reading with an  
independent meter and take corrective action if indicated.  
NOTE:  
RO membranes may achieve this level of performance under optimum conditions  
(good feed water).  
Alarm Set:  
The Alarm Set switch allows the operator to set the TDS levels at which the alarm  
will sound. The alarm should be set at twice the initial TDS reading. If the set  
point is exceeded, an audible alarm will sound.  
NOTE:  
Choose the alarm set point that is closest to twice the initial installation TDS  
reading.  
Alarm Mute:  
The Alarm Mute switch will mute the audible alarm for 30 - 45 seconds. A red  
indicator lamp is displayed on the switch when the mute switch has been pressed.  
12552e – 27 Feb 04  
1 - 10  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
1.9 Optional Equipment  
GE Osmonics offers a number of optional items to adapt the equipment to meet specific  
needs. Options are available to improve the quality of the feedwater to the RO.  
Frequently requested options are briefly described in the following paragraphs. If more  
information is needed, or if other options are desired, please consult the factory. Refer  
to Chapter xx for catalog part numbers.  
Carbon Filter Tanks remove chlorine/chloramine from the feed water, preventing  
damage to the RO membrane and injury to patients. The filter incorporates a bed of  
granular activated carbon. Two sizes of carbon tanks come in semi-automatic control  
valve and exchange head styles.  
WARNING:  
Carbon is not regenerated and must be replaced when exhausted.  
WARNING:  
A suitable carbon filter should always be present as pretreatment for the RO unit  
in order to prevent chlorine/chloramine damage to the thin film RO membrane  
and harm to the patient. GE Osmonics strongly recommends the use of two  
carbon tanks, used in a series configuration.  
Softeners remove calcium, magnesium, and other scale producing contaminants from  
the feed water. Their removal prevents the build-up of scale on the membrane surface.  
A control valve drives the regeneration of the softener, and will remove the hardness  
ions with a brine solution and rinse the ion exchange bed afterwards. The control valve  
is normally governed by a semi-automatic time clock. There are two sizes of softener  
tanks available.  
Multi-Media Filter Tanks are a multi layered filtration system that consists of various  
sized rocks as the media layers. The media are selected on the basis of their size and  
specific gravities. Capable of removing ninety-eight percent of all particles greater than  
ten microns, the multi-media filter is employed to remove suspended solids and  
particulates from the water. The use of the multi-media filter will extend the life of the  
pre-filter and RO membrane. Multi-media filters are not portable and not commonly  
used with portable RO systems.  
Dealkalizers are used to lower the pH of alkaline feed water. Water with pH levels  
above 8.5 with chloramines present can reduce the performance of the carbon filters  
and thin-film membranes. A control valve drives the regeneration rinsing of the  
dealkalizer. The control valve is normally governed by a semi-automatic time clock.  
There are two sizes of dealkalizer tanks available.  
12552e – 27 Feb 04  
1 - 11  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
1.10 Specifications:  
1.10.1 Input Water Requirements  
Minimum  
Maximum  
N/A  
Input flow Rate  
Temperature (water)  
NOTE:  
2.0 GPM  
10°C (50°F)  
32°C (90°F)  
Operating at lower temperatures may make it impossible for the F-801 RO to  
produce adequate permeate for the dialysis machine.  
pH (operating)  
TF Membranes  
5.5  
10.5  
pH w/chloramines present (operating)  
TF Membranes  
5.5  
8.5  
pH (cleaning)  
2.0  
11.0  
TF Membranes  
Pressure, dynamic  
30 PSI  
100 PSI  
1.10.2 Flow Rates  
Product Flow based on the following conditions:  
New Membranes  
Operating Pressure:  
Product Pressure:  
150 PSI  
1 PSI  
Feed Water Temperature:  
77°F (25°C)  
GPM LPM GPD LPD  
.41  
1.55 588  
2225.8  
Membrane flux may vary ±15%  
Waste Flow: 2000ml/min (±10%)  
1.10.3 Electrical Requirements  
Motor  
1/3 HP  
115 V~  
60  
Voltage  
Hertz  
Amps  
10  
Phase  
single  
8 ft (2.4 m)  
Power Cord (length)  
12552e – 27 Feb 04  
1 - 12  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
1.10.4 Membrane Performance Characteristics  
Salt rejection (TF Membranes):  
95% minimum at initial testing at factory.  
Inlet water 1000 PPM NaCl  
Test Solution:  
Conductivity:  
Operating Pressure:  
Product Back pressure:  
2000 micro mhos-cm  
200 PSI  
0 PSI  
1.10.5 Environmental Requirement  
Minimum  
Maximum  
32°C (90°F)  
Ambient Temperature  
Storage Temperature  
Altitude  
4°C (39°F)  
2°C (36°F)  
N/A  
32°C (90°F)  
10,000 feet  
CAUTION:  
This equipment must not be allowed to freeze. Irreparable harm to various  
components, including RO membranes may result.  
NOTE:  
The user/operator should recognize that moisture can be caused by condensation  
and is not necessarily an equipment leak. This equipment will function in the  
presence of condensation.  
1.11  
Disposal  
Disposal of this product or parts must be carried out according with local disposal  
codes.  
1.12  
Electromagnetic Interference  
This equipment can generate, uses and can radiate radio frequency energy and, if not  
installed and used in accordance with these instructions, may cause harmful  
interference to other devices in the vicinity. However, there is no guarantee that  
interference will not occur in a particular installation. If this equipment does cause  
harmful interference with other devices, which can be determined by turning the  
equipment off and on, the user is encouraged to try to correct the interference by one  
or more of the following measures:  
Reorient or relocate the receiving device  
Increase the separation between the equipment  
Connect the equipment into an outlet on a circuit different from that to  
which the other device(s) is connected  
Consult the field service technician or manufacturer for help.  
12552e – 27 Feb 04  
1 - 13  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
1.13 Service Assistance  
If service assistance is required, please take the following steps:  
Consult the “Troubleshooting” section of this manual (Chapter Five). If the problem  
cannot be identified and corrected by any of the procedures found in that section,  
then...  
Contact your Facility Equipment Technician. If the technician is unable to help then…  
Call GE Osmonics Technical Service Department at (800) 633-3080 or (206) 395-2200.  
Technicians are available for all calls between 7:00 a.m. and 4:30 p.m. PST, Monday  
through Friday. Technicians are also available at other times for emergency calls  
only. Product consultants will be on hand to discuss the problem with you and  
endeavor to rectify it over the phone. If the problem appears to be of a more serious  
nature, you will be given instructions regarding the action to be taken. Prior to making  
the phone call, you must be prepared to answer two questions:  
1. What RO do you have, i.e. F-801  
2. What is the serial number of your RO? (Label, found on back of Remote  
Monitor.)  
RO SERIAL NUMBER: ______________________  
12552e – 27 Feb 04  
1 - 14  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
1.14 Return Goods Authorization (RGA) Procedure  
If you wish to return goods for, warranty evaluation and/or credit please have your  
original sales order, invoice and device serial number available when you call GE  
Osmonics. Call GE Osmonics at (800)-633-3080 and request Technical Service. A GE  
Osmonics Technical Service representative will provide instructions and a return  
authorization number, which needs to be clearly written on the outside of the box used  
to ship your materials. All equipment must be shipped to GE Osmonics with the  
freight prepaid by the customer. Call our Customer Service Center with any questions  
or issues concerning freight claims and a representative will discuss your situation.  
All materials to be returned must be rendered into a non-hazardous condition prior to  
shipping.  
12552e – 27 Feb 04  
1 - 15  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
1.15 Symbols and Abbreviations  
“RENALIN®”  
Refers to Peracetic Acid and Peroxide-based Compounds, such as  
Renalin®, Minncare™, Peracidin™ (20% peroxide, 4% Peracetic  
Acid)  
ASTM  
AAMI  
CA  
American Society for Testing and Materials  
Association for the Advancement of Medical Instrumentation  
Cellulose Acetate  
C
Celsius  
CSA  
cc  
Canadian Standards Association  
Cubic Centimeters  
Cubic Foot (feet)  
cf  
cfu/ml.  
cm  
Colony Forming Units per milliliter  
Centimeters  
DI  
Deionization  
F
Fahrenheit  
FDA  
Ft  
Food and Drug Administration  
Foot (feet)  
GPD  
GPM  
GHT  
gr  
Gallons Per Day  
Gallons Per Minute  
Garden Hose Thread  
Grains  
GAC  
HZ  
Granular Activated Carbon  
Hertz  
HP  
Horse Power  
lbs.  
Pounds  
LED  
LPD  
LPM  
Kohm  
m.  
Light Emitting Diode  
Liters Per Day  
Liters Per Minute  
One Thousand Ohms  
Meter  
mg/L.  
Milligrams Per Liter  
12552e – 27 Feb 04  
1 - 16  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
ml.  
Milliliters  
MNPT  
MSDS  
Megohm/cm  
mw  
Male National Pipe Thread  
Material Safety Data Sheet  
Million Ohms/Centimeter  
Molecular Weight  
Sodium Chloride  
NaCl  
N/A  
Not Applicable  
NPT  
NSF  
National Pipe Thread  
National Sanitation Foundation  
Polyamide  
PA  
P/N  
Part Number  
ppm  
psi  
Parts Per Million  
Pounds Per Square Inch  
Polyvinyl Chloride  
Quick Disconnect  
Reverse Osmosis  
PVC  
QD  
RO  
ROMA  
TB  
Reverse Osmosis Membrane Assembly  
Terminal Barrier Strip: “TB2” (Example: “TB2-1” refers to the  
terminal marked number 1 on the Power Supply Terminal Barrier  
Strip)  
TF  
Thin Film  
TDS  
UL  
Total Dissolved Solids  
Underwriters Laboratory  
Volts Alternating Current  
Water for Injection  
Weight  
VAC  
WFI  
WT  
~
Volts Alternating Current  
12552e – 27 Feb 04  
1 - 17  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
NOTES:  
12552e – 27 Feb 04  
1 - 18  
General Information  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
F-801 Internal RO System  
CHAPTER 2:  
INSTALLATION GUIDELINES AND INITIAL SYSTEM START-UP  
12552e - 27 Feb 04  
2 - 1  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e - 27 Feb 04  
2 - 2  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
2.1  
Installation Considerations  
1.  
The RO unit requires 115 VAC power for operation and consumes 10 amps of  
current.  
2.  
The RO unit requires nominally 1.6 GPM water at 30 PSI. Warmer feed water  
(up to 25°) will produce more permeate, and will consume more water. The  
water may also require some pretreatment to ensure maximum performance and  
suitability. GE Osmonics recommends that feed water be chlorine/chloramine  
free and solids above 5 microns in size be filtered out.  
3.  
4.  
5.  
A drain with a minimum capacity of 1.6 GPM is required for the RO unit.  
Back flow prevention may be required in accordance with local codes.  
Special connections may be needed to connect to water supply. Connections  
vary with different facilities.  
2.2  
Installation Procedures:  
CAUTION:  
If a carbon tank is present on the RO system, the carbon tank must be completely  
rinsed of all carbon fines and ash before connecting to the RO unit. Otherwise,  
irreparable damage to the TF membrane will result (refer to Carbon Manual).  
WARNING:  
Do not rinse carbon tanks until they are ready to be used in service. Carbon tanks  
that have been wetted and allowed to sit unused will grow bacterial.  
WARNING:  
The ROMA is stored with 1 wt.% sodium metabisulfite from the factory. Follow  
all Material Safety Data Sheet (MSDS) guidelines that are included with the RO.  
CAUTION:  
Supply water with a pH above 8.5 and containing chloramines will adversely  
affect the operation of the RO membrane.  
1.  
For installation of the pretreatment cart and RO to the dialysis machine cabinet,  
and all tubing connections (except product to use), refer to the Installation  
Drawing section of the system manual, or contact GE Osmonics for Drawing  
p/n 1239784.  
2.  
3.  
4.  
Connect the product to use line to the product line rinse extension (GE  
Osmonics p/n 12393) and route to drain.  
Connect the feed water line to water supply and route the RO waste line to  
drain.  
Turn the water supply on and allow the water to flow through the system to  
drain for 3-5 minutes.  
12552e - 27 Feb 04  
2 - 3  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
NOTE:  
There will be little or no flow through the product line at this time.  
5.  
Make sure that the RO power switch is in the OFF position. Plug the power  
cord into a 115 VAC grounded outlet. Do not use an extension cord or an  
adapter.  
6.  
7.  
Ensure the dialyze/flush valve is in the dialyze position.  
Turn the RO power switch on and run the system for 30 minutes. Observe  
displays and note stabilization of rejection and TDS.  
8.  
After a minimum of 3 hours run time, disinfect and rinse the RO unit with an  
approved disinfectant before using for hemodialysis treatment (refer to Chapter  
Four for instructions on disinfection).  
WARNING:  
After disinfection, the product water must be verified free of residual disinfectant.  
WARNING:  
Ensure that TDS and membrane performance (% rejection) are normal for the  
regions water quality. Consult your facility physician. An AAMI analysis and  
bacteria test of the product water must be obtained and interpreted before using  
the RO unit for hemodialysis treatment (refer to the Product Water Analysis  
procedure and the Product Water Culture Procedure).  
CAUTION:  
New RO units with thin film (TF) membranes must be rinsed to drain with normal  
pressures for a minimum of three (3) hours before exposure to disinfectants such  
as Renalin and a minimum of six (6) hours before exposure to formaldehyde.  
NOTE:  
Ideally, rejection will be 95% or greater. If less than 95%, extend rinse period an  
additional 30 minutes.  
9.  
Close the product valve and remove the product rinse extension line. Connect  
product line to inlet line on the dialysis machine. Open the product valve.  
10.  
11.  
Start the dialysis machine into a rinse cycle.  
Set the product water TDS alarm limit switch as prescribed by the physician.  
The switch is normally set to twice the initial TDS reading (or “next higher”  
setting) at start-up.  
WARNING:  
The TDS alarm limit should not be moved from this setting unless instructed by a  
physician.  
12552e - 27 Feb 04  
2 - 4  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
12.  
13.  
Perform the following checks:  
Pump pressure gauge: between 140 – 160 PSI  
Pre-filter pressure differential: Both above 30 PSI, no greater difference  
than 8 PSI above clean filter readings.  
Product pressure: between 35-40 PSI.  
Percent rejection: greater than 95%, continue to rinse if lower.  
TDS at normal for region.  
The RO product may be used in conjunction with hemodialysis treatment  
equipment only after obtaining test results which validate compliance with  
excepted standards.  
AAMI Product Water Panel  
Microbiological Culture (colony count) results  
Endotoxin (pyrogen) level testing (if applicable)  
14.  
2.3  
The F-801 RO is now ready for use.  
Product Water Analysis Procedure:  
In order to ensure that the RO is performing at or above AAMI (Association for the  
Advancement of Medical Instrumentation) Standards, it is necessary to perform an  
AAMI Water Quality Analysis test. For new units, the RO must be operated (rinsed to  
drain) for at least two hours before taking a sample, to ensure that the membranes are  
free of preservative or other contaminants which would invalidate the test.  
WARNING:  
There are many documented clinical risks in using less than AAMI Standard  
product water (refer to Technote 103, "AAMI Hemodialysis Water Quality  
Standards").  
NOTE:  
Follow the procedures recommended by your water quality-testing laboratory if  
they vary from the following steps.  
Materials Required: Produce water sampling kit, sterile gauze, tape, and container  
(such as a small bucket).  
1.  
Ensure that all pre-RO devices are operating correctly and are performance  
verified by the appropriate testing.  
2.  
3.  
Connect the RO unit to the required utilities (feed water, drain, power).  
With the dialysis machine on, operate the RO unit with product and waste flows  
to drain for at least 15 minutes after the product water quality has stabilized  
before drawing the water sample.  
12552e - 27 Feb 04  
2 - 5  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.  
Close the product valve on the F-801, and disconnect the product line from the  
inlet water line of the dialysis machine. Tape a sterile gauze around the opening  
of the inlet water line to the dialysis machine.  
5.  
6.  
Hold the product line over the container and run product water for 60 seconds.  
Aseptically remove the lid from the sample bottle and rinse the bottle twice with  
RO product water.  
7.  
8.  
Fill the bottle completely with RO product water and cap it immediately.  
Fill out the laboratory requisition form. Ensure the requisition form has, at a  
minimum, information to identify the facility or location of the RO, the serial  
number or facility equipment number to ensure which RO was tested and name,  
phone number and address of whom to contact.  
9.  
Place the bottle and the completed requisition in the pre-addressed mailer and  
mail it directly to the laboratory.  
NOTE:  
Total chlorine must be measured on-site. Tests for Chlorine/Chloramine (total  
chlorine) are not typically reported on the AAMI Analysis. Chlorine and  
chloramine dissipate from the water sample over time and reported results will be  
inaccurate.  
12552e - 27 Feb 04  
2 - 6  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
NOTES:  
12552e - 27 Feb 04  
2 - 7  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e - 27 Feb 04  
2 - 8  
Installation Guidelines and Instructions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
F-801 Internal RO System  
CHAPTER 3:  
SYSTEM OPERATION  
12552e – 27 Feb 04  
3 - 1  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e – 27 Feb 04  
3 - 2  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
3.1 Daily Start-up  
WARNING:  
Verification of the absence of all sanitizers, cleaners, disinfectants and  
chlorine/chloramines must be performed before RO product water is used for any  
patient related uses.  
WARNING:  
Use of exhausted carbon in the presence of chlorine/chloramines will result in  
damage to the F-801 RO membrane and may cause serious patient injury.  
CAUTION:  
When moving the F-801 assembly, push only on the handle and in the direction of  
travel. The only exception to this should be when loading into elevators or going  
across thresholds when the gap is greater than 1” or the threshold is uneven. The  
F-801 should be pulled over these to prevent the machine from tipping over or  
causing damage to the casters.  
CAUTION:  
Do not push unit sideways as this may cause the unit to tip over. If it is necessary  
(such as in tight patient areas), use extreme caution and do not push it in a  
direction where it could fall onto someone.  
CAUTION:  
Do not try to catch the equipment if it starts to fall over. If moving up or down a  
ramp, get help if unable to handle the load.  
NOTE:  
Obtain a copy of the F-801 Operation Log Sheet (see Technote section of system  
manual). All the start-up data for the RO may be recorded on this log sheet.  
1.  
2.  
3.  
Connect the RO unit to the required utilities (feed water, drain, and power).  
Ensure that all the valves in the water supply line to the RO are open.  
Ensure that all pre-RO devices have been flushed and tested and are operating  
within their specifications.  
4.  
Ensure that the product line is connected to the inlet water supply line of the  
dialysis machine. Open the product valve located on the end of the product line.  
If the product valve is closed, no product water will be delivered to the dialysis  
machine.  
5.  
Turn on the water supply.  
12552e – 27 Feb 04  
3 - 3  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
WARNING:  
Only the cold water supply should be used. If blending of hot and cold water is to  
occur, an appropriate temperature blend valve must be used, and the temperature  
of the incoming water supply must be continuously monitored, and not be allowed  
to rise above 30°C (86°F). Excessively high water temperature may cause severe  
patient injury or death.  
6.  
Turn the F-801 RO unit ON. The on/off switch is located on the back of the  
remote monitoring module atop the dialysis machine.  
CAUTION:  
Some pump noise (chattering) is normal in the flush position. Excessive pump  
noise (chattering) can be damaging to the pump. The RO is in “flush” from the  
full open (flush) position to just before the fully closed (operate) position. Adjust  
it by partly closing the flush valve to reduce chattering of the pump.  
7.  
8.  
The "Percent Rejection" monitor should indicate 95% rejection within 2 minutes  
of start-up. It is normal for rejection to drop off until membranes have rinsed up  
to optimum quality.  
The TDS monitor should stabilize within 2 minutes of start-up. The TDS during  
this time may be higher than the alarm set point. The audible alarm may be  
muted at this time by pressing the alarm mute switch. The alarm mute switch  
will silence the audible alarm for 30 - 45 seconds each time the switch is used.  
WARNING:  
When the RO is used for dialysis service, a test for disinfectant residual MUST be  
performed.  
9.  
Perform the following routine checks before each use and record on the  
Operation Log sheet.  
WARNING:  
Any data that does not meet minimum standards set at your facility must be  
reported to the medical director. The RO should not be placed in service until  
readings are acceptable.  
Read the pump pressure gauge, it should read between 140 – 160 PSI.  
Read the inlet pre-filter pressure gauge, it should read between 30 – 100 PSI.  
Read the outlet pre-filter pressure gauge, it should read between 30 – 100 PSI  
with no more than 8 PSI difference between initial clean pre and post filter  
readings. Compare log sheet recordings for changes.  
Read the product pressure gauge, it should read between 35 – 40 PSI. The  
product pressure should be read with the product water hose connected to the  
Fresenius inlet, or with the product outlet valve closed.  
12552e – 27 Feb 04  
3 - 4  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
NOTE:  
If the product pressure or product flow is insufficient, a low dialysate flow alarm  
will occur on the dialysis machine. Small adjustments to the product pressure  
may be made with the pressure adjust knob located inside the RO module on the  
feed water line.  
Read the percent rejection. It should display in the green region between 85-  
95% rejection.  
WARNING:  
The product water must be shown to meet AAMI Standard Water Quality in the  
% rejection range displayed on that particular water source (refer to Technote  
103, AAMI Maximum Allowable Contaminant Level).  
Read the total dissolved solids (TDS). It should read normal for the region.  
The alarm should be set to twice the initial installation TDS reading or the next  
higher setting.  
WARNING:  
The product water must be shown to meet AAMI Standard Water Quality in the  
TDS range displayed on that particular water source (refer to AAMI Maximum  
Allowable Contaminant Level).  
NOTE:  
Mg/L may be used interchangeably with parts per million (PPM).  
NOTE:  
An alarm light is located to the right of the mute button that will illuminate if the  
product water total dissolved solids displays less than 0.5 mg/L. This is a rare  
phenomenon that can occur if the source or feed water is already highly purified;  
otherwise, the system needs to be checked.  
10.  
The F-801 RO is now ready for use. Proceed with normal set-up routine.  
12552e – 27 Feb 04  
3 - 5  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
3.2  
Bicarbonate Container Filling  
RO water may be dispensed from the product line of the F-801 Internal RO system for  
use in filling bicarbonate jugs, drawing water samples, etc.  
1.  
With the dialysis machine OFF and the F-801 ON and all tests and checks  
performed;  
a.  
Close the product valve on the product line.  
b.  
Disconnect the product line from the dialysis machine inlet water line.  
NOTE:  
GE Osmonics recommends to cover the inlet water line with sterile gauze and  
secure with tape. The inlet water line to the dialysis machine and the product line  
from the RO unit when disconnected, should always be treated as aseptically as  
possible. The inlet water line to the dialysis machine is not routinely disinfected,  
and is a prime place for bacteria to proliferate.  
c.  
Using aseptic technique, place the line over the bicarbonate container,  
open the product valve and dispense as needed.  
d.  
When finished dispensing, close the product valve, reconnect the product  
line to the inlet water line of the dialysis machine and open the product  
valve.  
2.  
Continue with your normal set-up routine.  
3.3  
Shut Down Procedure  
The following procedure, when performed in sequence, will significantly decrease the  
water pressure built-up in the RO system and prevent splashing.  
1.  
2.  
3.  
4.  
5.  
6.  
Turn OFF the dialysis machine.  
Turn OFF the F-801 RO unit.  
Turn OFF the inlet water supply, wait one minute.  
Disconnect the feed water line from the water source.  
Remove the drain line from the drain.  
Disconnect from the utilities and store the RO.  
NOTE:  
If the F-801 RO is to be stored for more than 24 hours (RO not being operated),  
the RO should be stored in an approved disinfectant or storage solution (Not  
‘RENALIN®’ or other type oxidizing products). Refer to the disinfection or  
storage procedure in this manual.  
12552e – 27 Feb 04  
3 - 6  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
NOTES:  
12552e – 27 Feb 04  
3 - 7  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e – 27 Feb 04  
3 - 8  
System Operation  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
F-801 Internal RO System  
CHAPTER FOUR:  
ROUTINE MAINTENANCE  
12552e – 27 Feb 04  
4- 1  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e – 27 Feb 04  
4- 2  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.1  
Routine Maintenance of F-801 Internal RO System  
GE Osmonics RO systems have been designed to operate with a minimum of operator  
attention. Like all mechanical systems they will operate longer and with less trouble  
when operator maintenance is performed regularly. Operator maintenance on a GE  
Osmonics reverse osmosis system is limited to maintaining performance logs, cleaning  
and disinfection, replacement of damaged or failed parts and periodic lubrication. The  
maintenance procedures have been categorized by their frequency of action and are as  
follows:  
WARNING:  
Prior to performing any electrical maintenance unplug the RO power cord from  
the wall receptacle.  
WARNING:  
Do not perform maintenance while RO is being used for patient treatment.  
CAUTION:  
Perform system disinfection as per established facility protocol and referring to  
the disinfection procedure in this manual.  
NOTE:  
The first part of this section provides the timelines and simple instructions for  
maintenance of your RO. The specific instructions and procedures for the more  
complex maintenance items are provided in the second part of this section.  
4.1.1 Daily/Per Treatment Maintenance  
1.  
Test and record the hardness level of the feed water in the pretreatment log.  
The hardness level should be less than two grains per gallon. Refer to  
appropriate pretreatment manual for detailed instructions.  
NOTE:  
Check the softener output for hardness level at the end of the operational day to  
confirm adequate softener capacity for next days treatment.  
2.  
3.  
Observe and record salt level in the brine tank (for systems with water softener  
pretreatment).  
Observe and record the Chlorine/Chloramine concentrations in the pretreatment  
log. Follow the instructions included with each test kit.  
WARNING:  
Take corrective action or contact your local service representative if any optional  
pretreatment device fails to provide adequate treatment of feed water before  
continuing with the use of the RO unit.  
12552e – 27 Feb 04  
4- 3  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.  
Observe and record the pre-filter inlet and outlet pressure gauges on the  
operation log. The difference in the pressure values will help measure the  
condition of the filter cartridge. When the pressure differential (delta P)  
increases 8 PSI above the clean, initial delta P, the cartridges need to be  
changed (or sooner if the pre-filter core appears soiled). Refer to section 4.2 for  
specific instructions on changing pre-filter.  
5.  
6.  
Observe and record the percent rejection and TDS water quality readings in the  
operation log.  
Observe and record the pump pressure reading. It should be between 140 – 160  
PSI. Adjustments may be made on the pump head. Refer to drawing section  
for details.  
7.  
Observe and record the product pressure reading. Reading should be 35 – 40  
PSI (with F-801 product valve closed or connected to dialysis unit but not  
operating). Adjustments to the product pressure may be made with the pressure  
adjust regulator knob located inside the RO module.  
NOTE:  
After adjusting product pressure, verify that pump pressure is within normal  
range.  
8.  
Measure and record the feed water temperature if blending hot and cold water.  
CAUTION:  
Damage may occur to the RO system membrane if the incoming water  
temperature is higher than 32°C (90°F).  
WARNING:  
Injury to the patient may result with water temperatures higher than 98°F.  
NOTE:  
Systems with blending valves usually have thermometers installed in the  
plumbing. Hand held thermometers may also be used. The water temperature has  
a significant effect on the flow rate of the RO membrane (refer to Technote 113  
Temperature Correction Factor chart) and on the evaluation of the membrane's  
need to be cleaned. Feed water temperature blending assemblies can be used on  
portable RO units, however GE Osmonics does not supply them.  
9.  
Disinfect the F-801 RO unit per established facility protocol (refer to  
Disinfection Procedure) if RO unit has sat dormant for longer than 24 hours or  
has been stored for long periods of time.  
12552e – 27 Feb 04  
4- 4  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.1.2 Weekly Maintenance  
1.  
Observe, or measure, and record the waste and product flow rates on the  
operation log. Allow the water temperature to stabilize before attempting to  
record the results. This will help prevent errors due to fluctuating temperature.  
a.  
b.  
Waste flow 2000 ml/min (±10%)  
Product flow 1580 ml/min (150 PSI pump pressure and 25C feed water  
temperature) with product line disconnected from dialysis machine and  
line open to atmosphere. Refer to Technote 113 for Temperature  
Correction Factors.  
2.  
3.  
Observe, or measure, and record the feed water temperature. Systems with  
blending valves have thermometers installed in the plumbing. A hand held  
thermometer may also be used. The water temperature has a significant effect  
in the product flow rate of the RO membrane and on the evaluation of the  
membranes’ need to be cleaned.  
Disinfect the system if needed. Refer to the disinfection procedure in this  
chapter.  
4.1.3 Monthly Maintenance  
1.  
Take a product water culture to test for bacteria colony forming units count  
(<200 cfu/ml). Refer to the product water culture procedure in this chapter.  
Disinfect if results indicate higher cfu count.  
NOTE:  
Water cultures should be taken prior to disinfection and after disinfection to  
determine effectiveness.  
2.  
Visually inspect the pre-filter core, if you are using cartridge pre-filtration.  
CAUTION:  
On occasion, the pressure differential between the two pre-filter gauges will not  
alter significantly, however, the pre-filter will be soiled. If the core of the pre-  
filter is dirty, particulates are breaking through the pre-filter and traveling  
downstream and causing damage to the pump and RO membranes. If this is the  
case, the pre-filter will require more frequent replacement on a routine basis.  
3.  
4.  
5.  
Check the pump pressure settings. The pressure should be 150 PSI ± 10 PSI,  
and should not exceed 200PSI.  
Check the product pressure setting. Adjust the pressure as needed with the  
regulator inside the cabinet.  
Perform trend analyses from the log sheets.  
12552e – 27 Feb 04  
4- 5  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.1.4 Quarterly Maintenance  
1.  
2.  
3.  
4.  
5.  
Clean the RO Membranes. Refer to the RO membrane cleaning procedure in  
this chapter.  
Disinfect the RO machine. The RO should be disinfected after cleaning. Refer  
to the membrane disinfection procedure in this chapter.  
Verify the TDS and percent rejection readings with an independent monitoring  
device every 3 months.  
Check the quick-disconnect fittings and hoses. They should seat securely and  
fluid should shut off when they are disconnected.  
Test feed water (tap water). Test for;  
a.  
b.  
c.  
d.  
Iron presence (critical when Renalin is used)  
pH alteration  
Chlorine / chloramines shifts  
Hardness fluctuations  
CAUTION:  
If feed water changes in its constituents, the pretreatment components may need to  
be reassessed for proper capacity.  
6.  
7.  
Inspect casters and caster mounting plates for damage. If damage is evident,  
replace casters.  
Tighten all hardware for the pretreatment stand assembly. This includes casters,  
handle bolts, leg attaching bolts, etc.  
4.1.5 Annual Maintenance  
1.  
An AAMI Water Quality Analysis of the product water should be conducted at  
least yearly. (Refer to the Water Quality Analysis Procedure in Chapter two.)  
2.  
3.  
4.  
5.  
6.  
7.  
8.  
Check the monitoring for accuracy.  
Check the gauges for correct functioning.  
Check the ball valves for correct operation.  
Check the tubing and fittings for leaks or wear.  
Check the switch and lights for correct operation.  
Check all pre RO devices for correct operation.  
Lubricate the motor. Refer to the lubricating instructions in this chapter.  
12552e – 27 Feb 04  
4- 6  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.2  
Specific Maintenance Procedures and Instructions  
4.2.1  
Motor Lubricant Procedure  
NOTE:  
Only perform this procedure on motors that have plugs on each end of the motor  
and have instructions on the motor identification plate indicating the requirement.  
NOTE:  
During initial start-up and after performing motor lubrication procedure, a hot  
oily smell may be detected. This is normal and should fade away.  
1.  
2.  
Remove the plastic plugs at both ends of the motor housing near the motor shaft.  
Follow instructions on motor identification plate for type and quantity of oil to  
be added. If unable to read motor plate, add 30 drops of SAE 20 wt. non-  
detergent motor oil into the opening.  
3.  
Replace the plastic plugs.  
Turn on the motor and run for 10 minutes.  
Leak repair  
4.  
4.2.2  
General information: Any leak from any part of the system should be corrected as soon  
as possible. Left unattended, leaks can cause secondary damage necessitating expensive  
repairs.  
NOTE:  
Use FDA/NSF approved silicone sealant or Teflon tape on all threaded  
connections.  
Threaded Connections:  
1.  
Prior to performing this maintenance on any threaded fitting, release the  
pressure on the piping/tubing system (i.e., turn off the pump, open a drain port,  
etc.)  
2.  
3.  
Remove the connecting tubing.  
Re-tape the pipe threads with 2 or 3 wraps of Teflon tape. (It is not necessary to  
remove the old Teflon tape, although it may be advisable to do so if tape is  
loose or in otherwise poor shape).  
4.  
5.  
Re-insert the threaded fitting into the opening and tighten hand tight. Be careful  
to avoid cross threading. Tighten further using a wrench only if necessary, and  
only about one-half turn.  
Re-connect piping, pressurize, and inspect for leaks.  
12552e – 27 Feb 04  
4- 7  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
6.  
7.  
If the leak is still present, de-pressurize system again and tighten the fitting  
another quarter-turn. Check again for leaks. (Over tightening can crack the  
female fitting).  
If the leak continues, it may be necessary to replace the male threaded fitting  
and/or the female threaded fitting.  
Tubing connections:  
12552e – 27 Feb 04  
4- 8  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
1.  
Prior to performing this maintenance on any tube fitting, release any pressure on  
the tubing system (i.e., turn off the pump, open a drain port, etc.).  
2.  
3.  
Remove the red locking clip (if applicable).  
Push the dark gray collar in towards the body of fitting (light gray). This  
releases the ‘gripper’, which holds the tubing in place.  
4.  
5.  
Hold the gray collar in while gently pulling the tube away from the fitting. It  
may be helpful to wiggle the tube side-to-side slightly while pulling.  
Remove the gray collar by pulling straight out. Reach into the opening with a  
small, blunt (non-scratching) probe and remove the O-ring. (Inspect the collar to  
insure that all grippers are intact.)  
6.  
7.  
Rinse the O-ring with warm running water to remove any dirt or debris and set  
aside to dry on a clean paper towel.  
Clean the inside of the fitting with a wet cotton swab or paper towel to remove  
any dirt or debris. Be sure not to leave any cotton fibers from the swab or towel  
inside the fitting.  
8.  
Lubricate the O-ring sparingly with silicone O-ring lubricant and place the O-  
ring back inside the fitting.  
9.  
Replace the gripper collar.  
10.  
With a tubing cutter or sharp utility knife, cut off approximately ½” to 1” of the  
tube. Carefully cut the tube as square (perpendicular to the length) as possible.  
Inspect remaining tubing for marks/scratches and replace/repair as necessary.  
11.  
12.  
Place the tube in the fitting opening and push firmly until the tube seats fully in  
the fitting (completely ‘bottomed out’).  
Pull back on the tubing to seat the grippers, and re-install the locking clip (if  
applicable) between the light gray body of the fitting and the dark gray collar  
(flat side towards the body).  
13.  
14.  
Re-pressurize the system (if applicable) and inspect for leaks.  
If the leak is not resolved, start over or simply replace the entire tube fitting or  
tube.  
4.2.3 Pre-filter Replacement  
1.  
Turn OFF the Dialysis machine and the F-801 RO system. Shut the feed water  
OFF.  
2.  
3.  
Ensure that the drain line is in the drain receptacle with the proper air gap.  
Slowly turn the blue pre-filter housing counter-clockwise, being careful to vent  
any remaining water pressure.  
12552e – 27 Feb 04  
4- 9  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.  
5.  
Carefully remove the housing (bottom part) from its head. Empty the water and  
used pre-filter cartridge.  
Inspect the central core of the used pre-filter for dirt or rust. If discoloration is  
present in the cartridge interior, the filter has been in service too long and  
should be replaced sooner, or at a delta pressure lower than 8 PSI.  
6.  
7.  
Check the O-ring in the groove of the filter housing. Replace the O-ring (GE  
Osmonics p/n 40694) if it appears worn or damaged.  
Clean and rinse the pre-filter housing.  
a.  
Optional; a 1% bleach solution may be used to sanitize the housing, rinse  
thoroughly.  
8.  
Insert the new pre-filter cartridge into the housing. Ensure that the pre-filter is  
firmly in place.  
9.  
Re-install the housing, tighten clockwise.  
10.  
Turn ON the feed water and check for leaks. This will also purge the air from  
the pre-filter housing.  
11.  
If there are no leaks, the RO system is ready for use.  
12552e – 27 Feb 04  
4- 10  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.3  
F-801 Cleaning Indications  
Foulants such as minerals, silt and organic substances may collect on the RO membrane  
surface and cause the RO membrane to foul and diminish in its function. Many of these  
foulants are prevented by proper selection of pretreatment equipment, while others are  
treatable only with more sophisticated designs. Regardless of the foulant, it is essential  
to clean the membrane promptly once cleaning is indicated. The longer a foulant is  
allowed to remain in contact with the RO membrane, the more difficult it will be to  
remove. In severe cases, repeated cleanings may be required.  
NOTE:  
Some foulants CANNOT be removed with cleaning and membrane performance  
may not be fully restored. Observe the appearance of the cleaning solution before  
discarding it. If the solution is heavily soiled, repeat the cleaning procedure with  
fresh solution of the same type (high pH or low pH).  
Routine cleaning of the RO membranes in conjunction with regular disinfection will  
keep bacteria counts at a minimum and the RO membrane performance at its best.  
Cleaning removes the silt, organic and mineral deposits that collect on the RO  
membrane surface which, if left untreated, encourage bacterial growth and will  
eventually impair the RO membranes' ability to perform. Disinfection kills most of the  
bacteria that may be present in the system, but it does not remove the deposits on the  
membrane.  
There are a variety of indications for cleaning membranes. The most common is a  
reduced product flow rate. Remember that a reduction in inlet water temperature can  
also result in a reduced flow rate. Refer to Technote 113 "Temperature Correction  
Factors" to determine the reduction in flow that can be attributed to reduced water  
temperature. Clean the membranes whenever the temperature compensated flow rate  
drops by 10%. A loss in the membrane performance (percent rejection) or increase in  
TDS of the product water may also be an indication to clean the membranes. A good  
preventative maintenance practice is to clean the membranes on a quarterly basis, and  
more often if needed.  
CAUTION:  
Membranes that have not been cleaned for a long time may actually incur damage  
during cleaning as perforations are exposed that were masked by accumulations of  
foulant. Regular maintenance cleaning is the best assurance of long membrane  
service.  
12552e – 27 Feb 04  
4- 11  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.3.1  
Pre-Cleaning Preparation  
There are two types of cleaners available, a low pH cleaner (Calcium/Iron removal) and  
a high pH cleaner (Silt/Organics removal). GE Osmonics provides these cleaners in  
liquid form. The low pH cleaner attacks mineral scale deposits such as calcium,  
magnesium and iron build-up, and should be used first. The high pH cleaner will  
remove accumulated silt and organic matter from the RO membrane.  
CAUTION:  
The different cleaners must not be mixed together. Introduce the low pH cleaner  
first, thoroughly rinse the system, and then use the high pH cleaner.  
WARNING:  
The solution is a mild irritant to eyes and skin. It is recommended that protective  
gloves and eye protection be worn when handling the cleaner. Neither respiratory  
protection nor special ventilation are required.  
Materials Required:  
LIQUICLEAN 103™ – (Low pH liquid) - GE Osmonics Part No. 1228653 (1 Gal)  
LIQUICLEAN 511™ – (High pH liquid) - GE Osmonics Part No. 1228791 (1 Gal)  
high and low pH cleaner  
2 Gallons pretreated water  
Disinfect container with volume of 2 ½ gallon or more  
pH test strips  
phosphate testing kit (optional)  
product rinse line (clear hose with male QD fitting on one end, open on the  
other)  
“DO NOT USE” warning labels (not provided by GE Osmonics)  
cleaning / disinfect log sheets  
most current log sheets for water quality comparison  
sterile gauze and tape  
4.3.1.1 Pre-Cleaning Procedure  
Read and understand instructions on cleaner container before beginning  
procedure.  
1.  
Connect the water inlet line to the water supply and the waste line to drain.  
Turn the RO unit and water supply on.  
2.  
Close the product valve. Disconnect the product line (white QD) from inlet line  
of the Fresenius dialysis machine. Cover exposed inlet water line with sterile  
gauze and secure with tape.  
3.  
Connect the product rinse line to the product line and direct to clean drain  
receptacle. Open the product valve.  
12552e – 27 Feb 04  
4- 12  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.  
5.  
Turn the RO on for five minutes and establish a pre-cleaning % rejection, TDS  
and pH record on the cleaning disinfect log sheet.  
Prepare the cleaning solution by adding the amount of cleaner as specified on  
the labeling to two gallons of pretreated or RO product water.  
CAUTION:  
Water to be used must be chlorine free or damage to the RO membranes may  
result.  
6.  
Turn the RO power switch off and close the product valve.  
WARNING:  
Label machine with appropriate warning signs such as "Do Not Use / Contains  
Cleaner" to prevent use of the RO until it is properly rinsed.  
7.  
8.  
9.  
Connect the product line to the white fitting on the disinfect/cleaning tank.  
Open the product valve.  
Connect the feed line (green QD) from the pre-filter outlet to the green fitting on  
the disinfect tank.  
Connect the drain line (red QD) extension and connect the female portion to the  
red fitting on the disinfect tank.  
CAUTION:  
By connecting all the lines to the disinfect/cleaning tank, the system is set-up for  
recirculation. Never leave the unit unattended in the recirculation mode.  
4.3.1.2 Cleaner Recirculation and Dwell Time  
1.  
Turn the Dialyze/Flush valve to the Flush position and open the product valve.  
The Flush mode will allow the flow of cleaner only to the rejection side of the  
RO membrane to flush surface particulates away. Very little or no product  
water will be made in flush mode.  
2.  
Turn the RO power switch ON. Let the RO operate for 5 to 10 minutes in the  
flush mode. Pump pressure during this step should read approximately 40 – 80  
PSI.  
CAUTION:  
Some pump noise (chattering) is normal in the flush position. Excessive pump  
noise (chattering) can be damaging to the pump. The RO is in “flush” from the  
full open (flush) position to just before the fully closed (dialyze) position. Adjust  
it by partly closing the flush valve to reduce chattering of the pump.  
CAUTION:  
DO NOT ALLOW PUMP TO RUN DRY. Damage will occur if pump is run  
dry.  
12552e – 27 Feb 04  
4- 13  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
CAUTION:  
Check that there is waste flow within 10 seconds. If not, turn off the RO and  
verify that the disinfect-line is connected properly.  
CAUTION:  
Do not allow air or foam to be drawn into the RO system, as this will damage the  
pump.  
NOTE:  
The % rejection will fall below 85% and the TDS will rise considerably when the  
cleaner is re-circulated through the F-801 RO unit.  
NOTE:  
The cleaners will foam; this is normal. If using a closed container, provide a vent  
to the atmosphere to relieve pressure.  
3.  
4.  
Verify that the F-801 and Fresenius dialysis unit are properly labeled with  
warning signs to prevent use during cleaning process.  
WARNING:  
The F-801 RO machine in now completely filled with cleaning solution and must  
NOT be used to supply water to the Fresenius dialysis machine or for any other  
purpose until a complete rinse procedure has been performed. See Cleaner Rinse  
Procedure below.  
Let the RO stand dormant with cleaning solution in it for sixty minutes with a  
maximum of overnight.  
4.3.1.3 Post-Cleaning Rinse Procedure  
1.  
2.  
After the dwell time, remove the waste hose and clear accessory flush line from  
the cleaning solution and place it in the drain with the appropriate air gap.  
Turn the RO on and run cleaning solution to the drain until the container is  
empty. This aids in emptying the container of cleaning solution.  
CAUTION:  
DO NOT ALLOW THE PUMP TO RUN DRY. As an alternative, the solution  
may be poured in the drain.  
CAUTION:  
Do not allow air or foam to be drawn into the RO system, as this will damage the  
pump.  
3.  
4.  
5.  
When the disinfect/cleaning tank is empty, turn the F-801 RO unit OFF.  
Reconnect the feed line to the pre-filter outlet (green QD).  
Open the product valve.  
12552e – 27 Feb 04  
4- 14  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
6.  
7.  
Ensure that the water supply to the RO unit is on. If it is partially closed, rotate  
the dialyze/flush valve to the full flush position.  
Turn ON the F-801 RO unit and operate in the flush mode for 5-10 minutes.  
This will aid in rinsing the system quickly.  
CAUTION:  
Some pump noise (chattering) is normal in the flush position. Excessive pump  
noise (chattering) can be damaging to the pump. The RO is in “flush” from the  
full open (flush) position to just before the fully closed (dialyze) position. Adjust  
it by partly closing the flush valve to reduce chattering of the pump.  
8.  
9.  
Rotate the dialyze/flush valve to the dialyze position.  
Turn the F-801 RO unit ON. Rinse for 20 minutes with the Dialyze/Flush valve  
in the Dialyze position until the % rejection and TDS monitors’ displays  
normalize (e.g. 'usual' % rejection and 'usual' TDS reading). Refer to your  
operating log for 'usual' readings and compare to the pre-cleaning values.  
A. During this rinse period, rotate the Dialyze/Flush valve to the  
Flush position briefly and then back to Dialyze three times to  
build-up and release pump pressure. This aids in flushing the  
cleaner from the system.  
10.  
11.  
Test the product water for post-cleaning pH and compare it to the pre-cleaning  
pH level. The values should be the same or very close to the pre-cleaning pH.  
If it is not, continue rinsing in the Dialyze position until equalization of the  
value occurs.  
When all of the parameters (% Rejection, TDS, pH) have normalized, operate  
the RO for an additional 20 minutes in rinse mode.  
NOTE:  
Repeat the cleaning procedure with the other pH cleaner. GE Osmonics  
recommends using a low pH cleaner first, followed by a high pH cleaner after  
rinsing thoroughly.  
12.  
13.  
Turn the F-801 RO unit OFF.  
Close the product valve and then disconnect the clear accessory flush tubing  
from the product line (white QD / valve) and store properly. Reconnect the  
product line to its original connection and open the product valve.  
14.  
Remove the “Do Not Use/Contains Cleaner” warning label.  
CAUTION:  
The F-801 RO unit must be disinfected after the cleaning procedure.  
12552e – 27 Feb 04  
4- 15  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.3.2  
BIOSAN Pre-Cleaning Preparation  
BIOSAN is a liquid cleaner/descaler agent intended for use in removing mineral scale  
and organic biofilm and is a short-term storage solution for RO water systems.  
WARNING:  
The solution is a mild irritant to eyes and skin. It is recommended that gloves  
(latex/vinyl) and eye protection be worn when handling BIOSAN. Neither  
respiratory protection nor special ventilation are required.  
CAUTION:  
BIOSAN is not intended for use in Hemodialysis Machines.  
NOTE:  
BIOSAN can be used as a low pH cleaner or as a short term (two weeks) storage  
solution.  
Materials Required:  
BIOSAN - GE Osmonics Part No. 14962 (1 Gal)  
product water  
Disinfect container with volume of 2 ½ gallon or more  
pH test strips  
product rinse line (extension hose)  
“DO NOT USE” warning labels (not provided by GE Osmonics)  
cleaning / disinfect log sheets  
most current log sheets for water quality comparison  
sterile gauze and tape  
4.3.2.1 BIOSAN Cleaning Procedure  
Read and understand all BIOSAN cleaning instructions before beginning  
procedure.  
1.  
Connect the water inlet line to the water supply, and the waste and product lines  
to drain. Turn the RO unit and water supply on.  
2.  
Establish the pre-cleaning % rejection, pH and TDS readings and record them  
on the Cleaning/Disinfect Log Sheet.  
3.  
4.  
Turn the RO OFF.  
Close the product valve. Disconnect the product line (white QD) from inlet line  
of the Fresenius dialysis machine. Cover exposed inlet water line with sterile  
gauze and secure with tape.  
5.  
Connect the product rinse line to the product line and direct to clean drain  
receptacle. Open the product valve.  
12552e – 27 Feb 04  
4- 16  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
6.  
7.  
Prepare the cleaning solution by adding the amount of cleaner as specified on  
the labeling to two gallons of pretreated or RO product water.  
CAUTION:  
Water to be used must be chlorine free or damage to the RO membranes may  
result.  
Turn the RO power switch off and close the product valve.  
WARNING:  
Label machine with appropriate warning signs such as "Do Not Use / Contains  
BIOSAN" to prevent use of the RO until it is properly rinsed.  
8.  
Connect the product line to the white fitting on the disinfect/cleaning tank.  
Open the product valve.  
9.  
Connect the feed line (green QD) from the pre-filter outlet to the green fitting on  
the disinfect tank.  
10.  
Connect the drain line (red QD) extension and connect the female portion to the  
red fitting on the disinfect tank.  
CAUTION:  
By connecting all the lines to the disinfect/cleaning tank, the system is set-up for  
recirculation. Never leave the unit unattended in the recirculation mode.  
4.3.2.2 BIOSAN Recirculation and Dwell Time  
1.  
Rotate the Dialyze/Flush valve to the Dialyze position. This will allow the  
BIOSAN to pass from the rejection side of the membrane through to the product  
side of the membrane for total system exposure to the cleaning solution once the  
RO is turned on in disinfect mode (step #2).  
2.  
Turn the F-801 RO unit ON. Let the RO operate for three minutes.  
CAUTION:  
Check that there is waste flow within 10 seconds. If not, turn RO OFF and verify  
that the disinfect line is connected properly.  
CAUTION:  
DO NOT ALLOW PUMP TO RUN DRY. Damage will occur if pump is run  
dry.  
CAUTION:  
Do not allow air or foam to be drawn up into the RO unit, as this will damage the  
pump.  
NOTE:  
BIOSAN will foam; this is normal. If using a closed container, provide a vent to  
the atmosphere to relieve pressure.  
12552e – 27 Feb 04  
4- 17  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
3.  
Turn the Dialyze/Flush valve to the Flush position and operate RO for an  
additional 5-10 minutes. The Flush mode will allow the flow of cleaner only to  
the reject side of the RO membrane to flush particles away. Very little or no  
product water will be made. Pump pressure during this step should read  
approximately 40-80 PSI.  
CAUTION:  
Some pump noise (chattering) is normal in the flush position. Excessive pump  
noise (chattering) can be damaging to the pump. The RO is in “flush” from the  
full open (flush) position to just before the fully closed (operate) position. Adjust  
by partly closing the flush valve to reduce the chattering of the pump.  
NOTE:  
The % rejection will fall below 85% and the TDS will rise considerably when the  
cleaner is re-circulated through the F-801 RO unit.  
4.  
5.  
Turn the F-801 RO unit OFF.  
WARNING:  
The F-801 RO machine in now completely filled with BIOSAN cleaning solution  
and must NOT be used to supply water to the Fresenius dialysis machine or for  
any other purpose until a complete rinse procedure has been performed. See  
Cleaner Rinse Procedure below.  
Allow the BIOSAN to dwell in the RO for a minimum of one hour.  
NOTE:  
Thin-film RO membranes: The F-801 RO unit may be stored in BIOSAN for up  
to two weeks.  
4.3.2.3 BIOSAN Post-Cleaning Rinse Procedure  
1.  
2.  
After the dwell time, remove the waste hose and clear accessory flush line from  
the cleaning solution and place it in the drain with the appropriate air gap.  
Turn the RO on and run cleaning solution to the drain until the container is  
empty. This aids in emptying the container of cleaning solution.  
CAUTION:  
DO NOT ALLOW THE PUMP TO RUN DRY. As an alternative, the solution  
may be poured in the drain.  
CAUTION:  
Do not allow air or foam to be drawn into the RO system, as this will damage the  
pump.  
3.  
4.  
When the disinfect/cleaning tank is empty, turn the F-801 RO unit OFF.  
Reconnect the feed line to the pre-filter outlet (green QD).  
12552e – 27 Feb 04  
4- 18  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
5.  
6.  
Open the product valve.  
Ensure that the water supply to the RO unit is on. If it is partially closed, rotate  
the dialyze/flush valve to the full flush position.  
7.  
Turn ON the F-801 RO unit and operate in the flush mode for 5-10 minutes.  
This will aid in rinsing the system quickly.  
CAUTION:  
Some pump noise (chattering) is normal in the flush position. Excessive pump  
noise (chattering) can be damaging to the pump. The RO is in “flush” from the  
full open (flush) position to just before the fully closed (dialyze) position. Adjust  
it by partly closing the flush valve to reduce chattering of the pump.  
8.  
9.  
Rotate the dialyze/flush valve to the dialyze position.  
Turn the F-801 RO unit ON. Rinse for 20 minutes with the Dialyze/Flush valve  
in the Dialyze position until the % rejection and TDS monitors’ displays  
normalize (e.g. 'usual' % rejection and 'usual' TDS reading). Refer to your  
operating log for 'usual' readings and compare to the pre-cleaning values.  
A. During this rinse period, rotate the Dialyze/Flush valve to the  
Flush position briefly and then back to Dialyze three times to  
build-up and release pump pressure. This aids in flushing the  
cleaner from the system.  
10.  
11.  
Test the product water for post-cleaning pH and compare it to the pre-cleaning  
pH level. The values should be the same or very close to the pre-cleaning pH.  
If it is not, continue rinsing in the Dialyze position until equalization of the  
value occurs.  
When all of the parameters (% Rejection, TDS, pH) have normalized, operate  
the RO for an additional 20 minutes in rinse mode.  
12.  
13.  
Turn the F-801 RO unit OFF.  
Close the product valve and then disconnect the clear accessory flush tubing  
from the product line (white QD / valve) and store properly. Reconnect the  
product line to its original connection and open the product valve.  
14.  
Remove the “Do Not Use/Contains BIOSAN” warning label.  
CAUTION:  
The F-801 RO unit must be disinfected after the cleaning procedure.  
12552e – 27 Feb 04  
4- 19  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.4  
F-801 Disinfection Indications  
DETERMINATION OF MICROBIAL CULTURING AND DISINFECTION  
FREQUENCY IS THE RESPONSIBILITY OF THE OPERATING FACILITY.  
The following is recommended for critical applications where bacteria and endotoxins  
are a major concern, such as for Hemodialysis.  
Disinfection should occur when bacterial growth is above the limits set by the facility,  
or when over 200 CFU/ml (maximum AAMI standard) is detected. Routine  
disinfection at lower CFU counts will inhibit colonization of bacteria in the system and  
the distribution plumbing.  
Rapid re-growth of bacteria following disinfection, or excessively high colony counts  
(i.e. TNTC), may indicate a need for more aggressive intervention. Strategies such as  
increased frequency of disinfection, longer dwell times, or alternative chemical agents  
may be employed.  
CAUTION:  
Glutaraldehyde and Actril™ are not recommended disinfectants for the F-801.  
4.4.1  
Formaldehyde Disinfection Preparation  
WARNING:  
Always wear rubber gloves, apron and facemask. Use proper ventilation during  
disinfection and rinse-out. A face respirator with an organic cartridge to remove  
formaldehyde fumes may be worn. DO NOT BREATHE FUMES OR ALLOW  
FORMALDEHYDE TO COME IN CONTACT WITH SKIN OR EYES. Refer to  
MSDS supplied by manufacturer of the formaldehyde or formalin product.  
WARNING:  
Label the machine with appropriate warning signs such as "Do Not Use / Contains  
Formaldehyde" to prevent use of the RO until it is properly rinsed.  
CAUTION:  
New replacement thin film membranes must be rinsed to drain with normal  
pressures for a minimum of six (6) hours before first exposure to formaldehyde,  
or irreversible damage may occur. No mandatory pre-rinse of the membranes is  
required for future disinfections.  
CAUTION:  
No sanitizing agent containing chlorine should be used to sanitize this system.  
Use of chlorine will damage membranes.  
12552e – 27 Feb 04  
4- 20  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
Materials Required:  
200 ml. Formaldehyde, 37%  
1 gallon pre treated or RO water  
Disinfect container with volume of 2 ½ gallon or more  
formaldehyde test kit  
product rinse line (extension hose)  
“DO NOT USE” warning labels or signs (not provided by GE Osmonics)  
cleaning / disinfect log sheets  
most current log sheets for water quality comparison  
proper protective body wear  
sterile gauze and tape  
4.4.1.1 Formaldehyde Pre-Disinfection Procedure  
Read and understand all formaldehyde disinfection instructions before beginning  
procedure.  
1.  
Connect the water inlet line to the water supply, and the waste and product lines  
to drain.  
2.  
Close the product valve. Disconnect the product line (white QD) from inlet line  
of the Fresenius dialysis machine. Cover exposed inlet water line with sterile  
gauze and secure with tape.  
3.  
4.  
5.  
Connect the product rinse line to the product line and connect the product line to  
the white fitting on the disinfect/cleaning tank. Open the product valve.  
Connect the feed line (green QD) from the pre-filter outlet to the green fitting on  
the disinfect tank.  
Connect the drain line (red QD) extension and connect the female portion to the  
red fitting on the disinfect tank.  
CAUTION:  
By connecting all the lines to the disinfect/cleaning tank, the system is set-up for  
recirculation. Never leave the unit unattended in the recirculation mode.  
NOTE:  
Follow your facility’s procedure for formaldehyde disposal. The drain line may  
have to be placed in a container. Verify that the operate/flush valve is in the  
OPERATE position.  
12552e – 27 Feb 04  
4- 21  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
CAUTION:  
Formaldehyde or Formalin is usually provided in a clinical setting in a 37%  
solution. Check the labeling on the packaging to make certain this is the case. If  
the solution is other than 37%, the dilution will require adjustment. The end  
solution should not exceed 2.2%. Consult the Medical Director or Director of  
Nursing to determine the correct solution strength. GE Osmonics will not be  
responsible for any damaged components due to incorrect solution strength.  
4.4.1.2 Formaldehyde Disinfection and Dwell Time  
1.  
Rotate the Dialyze/Flush valve to the Dialyze position. This will allow the  
formaldehyde to pass from the rejection side of the membrane through to the  
product side of the membrane for total system exposure to the cleaning solution  
once the RO is turned on in disinfect mode (step #2).  
2.  
Turn the F-801 RO unit ON. Let the RO operate for 5-10 minutes. Product  
pressure during this step should read near zero.  
CAUTION:  
Check that there is waste flow within 10 seconds. If not, turn RO OFF and verify  
that the disinfect line is connected properly.  
CAUTION:  
DO NOT ALLOW PUMP TO RUN DRY. Damage will occur if pump is run  
dry.  
CAUTION:  
Do not allow air or foam to be drawn up into the RO unit, as this will damage the  
pump.  
3.  
4.  
Before the disinfect reservoir is emptied, collect a 10mL sample of solution  
from the product rinse line to be tested for the presence of formaldehyde. This  
test should indicate the presence of formaldehyde.  
WARNING:  
Ensure the machine is labeled with appropriate warning signs such as "Do Not  
Use / Contains Formaldehyde" to prevent use of the RO until it is properly rinsed.  
Turn off the F-801 RO and allow to dwell for 24 hours (4 hours minimum).  
NOTE:  
The RO may be stored for up to one year at this time. If this is the case, a 0.5%  
solution may be used instead of 2%.  
12552e – 27 Feb 04  
4- 22  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.4.1.3 Formaldehyde Post-Disinfection Rinse Procedure  
1.  
2.  
3.  
After the dwell time, remove the waste hose and clear accessory flush line from  
the disinfect solution and place it in the drain with the appropriate air gap.  
Connect the feed line to the pre-filter outlet.  
Connect the feed line to the water supply. Ensure that the water supply to the  
RO unit is ON. If it is partially closed, rotate the dialyze/flush valve to the full  
flush position.  
Turn ON the F-801 RO unit and operate in the flush mode for 2-3 minutes.  
This will aid in rinsing the system quickly.  
CAUTION:  
Some pump noise (chattering) is normal in the flush position. Excessive pump  
noise (chattering) can be damaging to the pump. The RO is in “flush” from the  
full open (flush) position to just before the fully closed (dialyze) position. Adjust  
it by partly closing the flush valve to reduce chattering of the pump.  
3.  
4.  
Rotate the dialyze/flush valve to the dialyze position.  
Turn the F-801 RO unit ON. Rinse for 20 minutes with the Dialyze/Flush valve  
in the Dialyze position until the % rejection and TDS monitors’ displays  
normalize (e.g. 'usual' % rejection and 'usual' TDS reading). Refer to your  
operating log for 'usual' readings and compare to the pre-cleaning values.  
A. During this rinse period, rotate the Dialyze/Flush valve to the Flush  
position briefly and then back to Dialyze three times to build-up and  
release pump pressure. This aids in flushing the cleaner from the  
system.  
5.  
Test product water for formaldehyde with test kit. If the test is positive,  
continue to rinse until a negative test is obtained.  
6.  
7.  
When test for formaldehyde is negative, turn the F-801 RO unit OFF.  
Close the product valve and then disconnect the clear accessory flush tubing  
from the product line (white QD / valve) and store properly. Reconnect the  
product line to its original connection and open the product valve.  
8.  
Remove the “Do Not Use/Contains Formaldehyde” warning label.  
9.  
The F-801 Internal RO system is now ready for use.  
NOTE:  
Success of the disinfection procedure may be ascertained by taking a water culture  
after completion. See the Product Water Culture Procedure in this chapter.  
12552e – 27 Feb 04  
4- 23  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.4.2  
Peracetic Acid/Peroxide-Based Compounds (Renalin®) Disinfection  
Procedure  
Renalin®, Minncare™, Peracidin™ and others are trade names for compounds based on  
Peracetic Acid and Hydrogen Peroxide compounds, and are common in dialysis  
facilities. The products referred to in this procedure typically have 20% hydrogen  
peroxide and 4% peroxacetic acid before dilution. Other products are available with  
these same chemicals in different ratios; these are not acceptable for use in the F-801  
RO. Actril™ and other compounds with different combinations of these chemicals  
must not be used on polyamide (thin-film) membranes. The name, “Renalin®” will  
be used in this manual to refer to acceptable products.  
WARNING:  
Follow all Material Safety Data Sheet (MSDS) precautions and OSHA Standards.  
CAUTION:  
New replacement thin film membranes must be rinsed to drain with normal  
pressures for a minimum of three (3) hours before first exposure to Renalin. No  
mandatory pre-rinse of the membranes is required for future disinfections.  
CAUTION:  
‘RENALIN®’ is corrosive to most metals other than stainless steel. The F-801 RO  
uses only compatible materials. ‘RENALIN®’ type products contain oxidizing  
agents and are potentially damaging to thin film (polyamide) membranes. It is  
imperative that all the precautions and instructions are followed correctly. Failure  
to comply with these instructions, which incorporate instructions from the  
membrane manufacturer, will jeopardize the system performance and the  
warranty.  
CAUTION:  
No sanitizing agent containing chlorine should be used to sanitize this system.  
Use of chlorine will damage membranes.  
Materials Required:  
correct amount of disinfectant, per manufacturers’ instructions  
1.5 gallons coldest RO water  
Disinfect container with volume of 2 ½ gallon or more  
product rinse line (extension hose)  
“DO NOT USE” warning labels or signs (not provided by GE Osmonics)  
cleaning / disinfect log sheets  
most current log sheets for water quality comparison  
sterile gauze and tape  
gross & residual disinfectant test strips  
proper protective body wear  
12552e – 27 Feb 04  
4- 24  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.4.2.1Renalin® Pre-Disinfection Preparation  
CAUTION:  
When ‘RENALIN®’ is to be used as a disinfectant, several preparatory steps are  
required by GE Osmonics as well as the membrane manufacturer. Failure to  
comply with these instructions, which incorporate instructions from the membrane  
manufacturer, will jeopardize the system performance and warranty.  
a. The system must be free of all incompatible materials.  
b. Clean the membranes with an acid such as 0.1% HCl or 0.4% phosphoric  
acids. Calcium and Iron Cleaner may be used as an alternate. It is essential  
that all iron and other transition metals be removed from the membrane  
surface.  
CAUTION:  
The RO system must be disconnected from the tempered water source (if present).  
The solution temperature must be monitored closely. At temperatures above 25°  
the ‘RENALIN®’ solution becomes very aggressive towards the RO membranes  
and will damage them in a short period of time. The disinfection procedure will  
raise the temperature of the solution. The operator must be prepared to add RO  
ice to the solution or discontinue the procedure when the water temperature  
approaches 77ºF /25°C.  
NOTE:  
At a minimum, a low pH cleaning should be performed prior to disinfection. It is  
preferred that a low and high pH cleaning be done.  
1.  
Read and understand the ‘RENALIN®’ disinfection procedure. Pay particular  
attention to steps a and b in CAUTION statement above before initiating the  
procedure.  
2.  
3.  
Ensure the cleaning/disinfectant container is clean. Add 1-1/2 gallons (5.7  
liters) of RO permeate (product water) to the container.  
Add disinfectant concentrate per manufacturer’s recommendation to the  
container and mix to prepare a 1:100 (1%) dilution.  
WARNING:  
Label the machine with appropriate warning signs such as "Do Not Use/ Contains  
Disinfectant” to prevent use of the RO until properly rinsed.  
4.  
5.  
Close the product valve. Disconnect the product line (white QD) from inlet line  
of the Fresenius dialysis machine. Cover exposed inlet water line with sterile  
gauze and secure with tape.  
Connect the product rinse line to the product line and connect the product line to  
the white fitting on the disinfect/cleaning tank. Open the product valve.  
12552e – 27 Feb 04  
4- 25  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
6.  
7.  
Connect the feed line (green QD) from the pre-filter outlet to the green fitting on  
the disinfect tank.  
Connect the drain line (red QD) extension and connect the female portion to the  
red fitting on the disinfect tank.  
CAUTION:  
By connecting all the lines to the disinfect/cleaning tank, the system is set-up for  
recirculation. Never leave the unit unattended in the recirculation mode.  
4.4.2.2Renalin® Disinfection and Dwell Time  
1.  
2.  
Verify that the dialyze/flush valve is in the DIALYZE position.  
Turn the RO ON and circulate the solution through the system for five minutes.  
After three minutes of circulation, test the product line for the presence of  
disinfectant. Test by performing the following steps. If test results are  
negative, continue circulating until a positive test is achieved.  
a. Close the product valve.  
b. Remove product line from tank.  
c. Hold line over sample container and slowly open the product valve.  
d. Test solution with a “gross” presence disinfectant test strip.  
e. If positive, close the product valve, return line to tank, and open product  
valve.  
WARNING:  
If a positive PRESENCE test result is not achieved, continue to recirculate for an  
additional 3-5 minutes, then perform presence test again. Recirculate and re-test  
as necessary until a positive test is obtained.  
CAUTION:  
Check that there is waste flow within 10 seconds. If not, turn RO OFF and verify  
that the disinfect line is connected properly.  
CAUTION:  
DO NOT ALLOW PUMP TO RUN DRY. Damage will occur if pump is run  
dry.  
CAUTION:  
Do not allow air or foam to be drawn up into the RO unit, as this will damage the  
pump.  
CAUTION:  
Do not allow the solution temperature to climb above 25°C during this step.  
Failure to control temperature will result in rapid deterioration of the membrane,  
resulting in poor rejection.  
12552e – 27 Feb 04  
4- 26  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
3.  
Turn off the RO; do not operate the RO dry.  
CAUTION:  
Do not allow air to be drawn up into the RO unit as this will damage the pump.  
4.  
5.  
Disconnect the drain line from the tank and re-attach the drain line extension.  
SECURE the drain line extension in the drain. Make sure the RED QD  
connector is fully joined and latched (listen for the ‘click’).  
CLOSE the ‘Product Valve’. Disconnect the product line from the tank, and  
attach the product rinse line extension’ to the product line. The Product Valve  
should remain closed at this time.  
WARNING:  
Make sure the drain line is secured in the drain. Disinfectant will be flowing out  
of the drain line during this step. Turn the RO unit power switch ON and observe  
the disinfectant level in the tank. It will begin to empty to drain. Turn the power  
switch OFF when the disinfectant tank is almost empty.  
CAUTION:  
Do not allow air to be drawn up into the RO unit as this will damage the pump.  
6.  
7.  
Disconnect the feed water line (green QD) from the disinfect tank, and  
disconnect the RO drain line extension (red QD). This closes the internal valves  
in the QD connectors, sealing the lines to prevent leakage of disinfectant.  
Allow the RO to soak in the disinfecting solution for 36 minutes -12 hrs.  
Thirty-six minutes contact time is expected to result in 90% bacteria kill; 12  
hours is expected to kill 99% of the bacteria. The minimum amount of soak to  
achieve disinfection is better for the membranes.  
CAUTION:  
Renalinmay not be left in contact with RO membrane for longer than 12 hours  
or damage to the RO membrane will occur. Do not store the unit in an overly  
warm room such as a utility room with a water heater, or boiler room during  
dwell time.  
8.  
Ensure that “Do Not Use / Contains Disinfectant” signs are displayed in  
appropriate places.  
CAUTION:  
The RO system is now filled with the disinfectant solution. THE USER MUST  
ENSURE THAT THE MACHINE IS NOT USED to provide water for any  
purpose whatsoever until the unit has been thoroughly rinsed and tested to verify  
absence of residual disinfectant in the product water stream.  
12552e – 27 Feb 04  
4- 27  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.4.2.3Renalin® Post-Disinfection Rinse Procedure  
1.  
After the dwell time, ensure that the drain line is in the drain with appropriate  
air gap, the feed line is connected to the pre-filter outlet, and that the clear  
accessory flush line is connected to the product line and placed in the drain  
receptacle with the appropriate air gap.  
2.  
3.  
Open the product valve.  
Connect the feed line to the water supply. Turn ON the water supply.  
CAUTION:  
Ensure that there is flow to drain.  
4.  
5.  
Rotate the dialyze/flush valve to the flush position.  
Turn the F-801 RO unit ON. Rinse for 20 minutes with the Dialyze/Flush valve  
in the Dialyze position until the % rejection and TDS monitors’ displays  
normalize (e.g. 'usual' % rejection and 'usual' TDS reading). Refer to your  
operating log for 'usual' readings and compare to the pre-cleaning values.  
A. During this rinse period, rotate the Dialyze/Flush valve to the Flush  
position briefly and then back to Dialyze three times to build-up and  
release pump pressure. This aids in flushing the cleaner from the  
system.  
6.  
Test product water for disinfectant with test kit. If the test is positive, continue  
to rinse until a negative test is obtained.  
7.  
8.  
When test for disinfectant is negative, turn the F-801 RO unit OFF.  
Close the product valve and then disconnect the clear accessory flush tubing  
from the product line (white QD / valve) and store properly. Reconnect the  
product line to its original connection and open the product valve.  
8.  
Remove the “Do Not Use/Contains Disinfectant” warning label.  
9.  
The F-801 Internal RO system is now ready for use.  
NOTE:  
Success of the disinfection procedure may be ascertained by taking a water culture  
after completion. See the Product Water Culture Procedure in this chapter.  
12552e – 27 Feb 04  
4- 28  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.5  
Product Water Culture  
FREQUENCY: Once a month or more often, depending upon culture results, and unit  
protocol. We recommend before and after disinfection.  
Materials Required:  
Bucket or similar container  
Sterile specimen container, such as Millipore (red) or 3M Commercial Standard  
Plate count sampler, or a container your lab may require  
RATIONALE: This procedure will help ensure that the microbial count of the product  
(RO) water falls within the AAMI Hemodialysis Water Standard. The AAMI standard  
maximum allowable concentration for bacteria in product water is 200 colony-forming  
units. A bacterial count above the maximum level may be harmful to the hemodialysis  
patient.  
OTHER: Consult your lab to find out what type of sterile container to use and how  
much product (RO) water is required to perform the test. Make certain that your  
laboratory is informed of the correct procedure for performing a product water culture.  
A "spread plate method" NOT a "calibrated loop" technique should be used, with  
tryptic soy agar or equivalent agar, NOT blood or chocolate agar. If tested incorrectly,  
the results may be inaccurate. The sample should be assayed within 30 minutes of  
obtaining it, or refrigerated (5°C) immediately and assayed within a 24 hour period.  
Obtain the samples wearing long sleeves and a mask to prevent contamination of  
samples, use a “mid-stream” “clean catch” type procedure.  
12552e – 27 Feb 04  
4- 29  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.5.1 Product Water Culture Procedure  
This procedure assumes the RO is connected to power and water, and is in operating  
condition.  
1.  
Close the product valve and connect the extension hose to the product line. .  
Place the other end in a drain and open the product valve.  
2.  
3.  
4.  
Turn on the RO unit. Allow the RO unit to run for 10-15 minutes.  
Close the product valve on the product line and remove the extension hose.  
Aseptically clean inside and outside of the product line quick disconnect fitting  
with three Betadyne swabs (optional).  
5.  
Using aseptic technique, hold the product line over a container, and allow  
approximately 100 cc of product (RO) water to flow to the drain (or allow  
product water to flow for 30 seconds).  
6.  
7.  
Maintaining aseptic technique, hold the product line over the sterile container  
with the product water flowing.  
Fill the container with RO product water. (If using commercial samplers, follow  
the manufacturer's instructions.)  
8.  
9.  
Aseptically cap the specimen container immediately.  
Label the specimen appropriately with:  
Test to be performed - "culture / colony count"  
Sample source - product water/RO, machine serial number and room  
location  
Time and date sample obtained  
Person who obtained specimen  
Any other pertinent information or procedures your facility or lab requires  
10.  
The samples should be assayed within 30 minutes of obtaining or refrigerated at  
5° C immediately and assayed within a 24-hour period.  
12552e – 27 Feb 04  
4- 30  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.6  
System Storage  
This storage procedure has been established to store F-801 RO machines. It is not  
recommended that a unit be dormant longer than 24 hours without disinfectant. ROs  
that have been left unused for more than 24 hours risk the possibility of colonizing  
bacteria and exceeding acceptable limits. ROs to be stored should first be disinfected  
with one of the procedures in this chapter. The RO should be stored in a location where  
it cannot freeze or become overly hot.  
CAUTION:  
‘RENALIN®’ products (Renalin®, Minncare™, Peracidin™. etc.) are not  
acceptable solutions for storage of systems. Exposure of more than 12 hours to  
RENALIN® will result in irreversible damage to the membranes.  
4.6.1 Formaldehyde Storage Procedure  
WARNING:  
Always wear rubber gloves, an apron and a facemask. If the room has ventilation,  
use it. If not, wear a face respirator with an organic cartridge to remove  
formaldehyde fumes. DO NOT BREATHE FUMES OR ALLOW  
FORMALDEHYDE TO COME IN CONTACT WITH SKIN OR EYES Refer to  
the MSDS supplied by the manufacturer of the formaldehyde/formalin product.  
1.  
If formaldehyde is chosen to disinfect the RO, follow the instructions for  
formaldehyde disinfection.  
2.  
3.  
Do not rinse out after the disinfectant has been circulated in the machine.  
Disconnect the disinfect line, the feed water line, the waste line, power and the  
product line.  
4.  
Coil the hoses and power cord, and cover the RO with a large plastic bag to  
protect the RO from dust.  
WARNING:  
After storage, the RO unit must be completely rinsed of formaldehyde, and a  
bacterial culture performed and interpreted before use. Perform AAMI analysis if  
it has been one year or more since last done.  
NOTE:  
All ROs can be safely stored in 0.5 to 2 % formaldehyde for up to one year.  
12552e – 27 Feb 04  
4- 31  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.6.2 BIOSAN Storage Procedure  
WARNING:  
It is recommended that gloves (latex/vinyl) and eye protection be worn when  
handling BIOSAN.  
1.  
2.  
3.  
Follow the BIOSAN procedure in this chapter.  
Do not rinse the RO out after BIOSAN has been circulated in the RO.  
Disconnect the disinfect line, waste line, feed water line, power and product  
line.  
4.  
Coil the hoses and power cord, and cover the RO with a large plastic bag to  
protect the RO from dust.  
WARNING:  
After storage, the RO unit must be completely rinsed of BIOSAN, and a bacterial  
culture performed and interpreted before use. Disinfect RO unit if necessary.  
Perform AAMI analysis if it has been one year or more since last done.  
CAUTION:  
The F-801 RO may be stored for up to two weeks in BIOSAN. After two weeks,  
if longer storage is needed, the RO may be rinsed and re-packed with BIOSAN.  
12552e – 27 Feb 04  
4- 32  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
NOTES:  
12552e – 27 Feb 04  
4- 33  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e – 27 Feb 04  
4- 34  
Routine Maintenance  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
F-801 Internal RO System  
CHAPTER FIVE:  
TROUBLESHOOTING  
12552e - 27 Feb 04  
5- 1  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e - 27 Feb 04  
5- 2  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
Table of Contents  
SECTION  
5.1  
TITLE  
PAGE  
5-5  
RO Fails to Run  
5.2  
RO Runs Intermittently  
Motor Runs Hot  
5-5  
5.3  
5-6  
5.4  
Low Pump Pressure  
5-6  
5.5  
Pressure Gauge Vibrates Erratically  
Excess Product Pressure (too high)  
Product Flow Too Low  
Product Flow Too High  
Membrane Product Pressure too Low  
Leaks  
5-7  
5.6  
5-7  
5.7  
5-7  
5.8  
5-8  
5.9  
5-9  
5.10  
5.11  
5.12  
5.13  
5.14  
5.15  
5-9  
Low Percent Rejection  
5-9  
Rejection Display Fluctuates  
TDS Value High  
5-10  
5-10  
5-11  
5-12  
RO Fails to Draw Disinfectant  
RO Pump Leaks  
12552e - 27 Feb 04  
5- 3  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e - 27 Feb 04  
5- 4  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
5.1  
RO Fails to Run  
1.  
Power off.  
a. Check position of power switch.  
b. Check power cord (plugged in).  
c. Check outlet and circuit breaker/fuse panel for power with voltmeter.  
d. Check fuses in RO monitor panel.  
2.  
Improper line voltage.  
a. Verify that line and equipment voltages are compatible.  
b. Check to make sure that combination of RO and other equipment that may be  
plugged into the same outlet are not overloading the circuit.  
3.  
4.  
Loose electrical connections.  
a. Check all wiring connections to verify solid connection.  
b. Verify that plug is securely connected to outlet.  
Pump motor inoperative.  
a. Verify that voltage supply is correct and motor is wired for correct voltage.  
Correct as necessary.  
b. Motor thermal overload switch tripped (most motors have automatic internal  
reset). Motor will restart when it cools approximately 10-15 minutes. Correct  
reason for the overload.  
c. Remove pump head and try restarting motor.  
d. Replace motor.  
5.  
Pump inoperative.  
a. Inspect pump and motor coupling for wear.  
b. Clean any debris or liquid from connection area.  
c. Replace pump.  
d. See 5.4.  
5.2  
RO Runs Intermittently  
1.  
Loose electrical connection.  
a. Verify that all wiring connections are secure.  
b. Verify that plug is securely connected to outlet.  
2.  
3.  
Pump pressure too high.  
a. Adjust pressure at pump head.  
b. Check waste flow-path for obstruction.  
c. See 5.1.4  
d. See 5.1.5  
Thermal overload trips  
a. Verify line voltage and motor amperage draw.  
b. Disassemble pump/motor at clamp and inspect coupler for wear.  
c. Pump pressure too high (see 5.2.2).  
12552e - 27 Feb 04  
5- 5  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.  
Pump inoperative  
a. Inspect pump and motor coupling for wear.  
b. Clean any debris or liquid from connection area.  
5.3  
Motor Runs Hot  
NOTE:  
Under normal conditions, the temperature of the motor may operate up to 40C  
(104F) above ambient (room) temperature.  
1.  
Pump/motor binding.  
a. Remove pump head and try restarting motor.  
b. Inspect pump and motor coupling for wear.  
c. Clean any debris or liquid from connection area.  
d. See 5.4  
e. Replace pump.  
2.  
3.  
Excessive current draw.  
a. Verify that flow rates and pressures are within RO specifications.  
b. Check motor/pump coupling for alignment.  
c. Confirm amperage draw with meter.  
d. Motor insulation shorted, replace motor.  
Voltage too low.  
a. Check voltage against serial label and motor label, correct condition.  
5.4  
Low Pump Pressure  
1.  
No pressure develops above level of feed pressure.  
a. Verify motor operation and water line connections.  
b. Verify flush valve is in the dialyze position.  
c. Verify (lower than normal) flow from product, if high see membrane problem.  
d. Verify (lower than normal) flow from waste, if high check flush valve  
operation.  
e. Pump damaged or worn out, replace pump.  
2.  
Pressure develops but does not achieve normal operating pressure.  
A. Waste or product flow too high.  
a. Check water temperature and flow rate.  
b. Check quality of product water.  
B. Faulty inlet pressure regulator.  
a. Replace regulator.  
C. Pump adjusted improperly  
a. Adjust pump at pump head and regulator.  
D. Pump worn out or damaged.  
a. Replace pump.  
12552e - 27 Feb 04  
5- 6  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
E. Flush valve is partly open  
a. Open and close valve and then measure and compare to original  
specifications.  
F. Pre-filter is plugged  
a. Replace pre-filter.  
5.5  
Pressure Gauge Vibrates Erratically  
1.  
Air in gauge tubing.  
a. Loosen tube fitting at gauge to bleed air.  
2.  
3.  
4.  
Obstruction in pressure gauge orifice.  
a. Remove obstruction.  
Pump telegraphing (making noise)  
a. Inspect pump and motor coupling for wear.  
Faulty gauge.  
a. Replace gauge.  
5.6  
Excess Product Pressure (too high)  
1.  
Inlet pressure regulator improperly adjusted.  
a. Adjust pressure regulator to correct product setting.  
2.  
Product flow rate too high.  
a. Decrease pump operating pressure if above specifications.  
b. Feed water temperature too high. Reduce temperature or discontinue use of RO  
to prevent damage to membranes.  
c. Damage or failure of membrane has occurred, replace membranes.  
5.7  
Product Flow Too Low  
Low water temperatures, fouled membrane elements, or inadequate pump pressure can  
cause low product flow. A partial blockage of the feed stream (dirty pre-filter) can also  
result in low product water flow by causing the pump to draw a strong vacuum.  
Similarly, an obstruction (defective check valve) in the product flow path may cause  
reduced flow.  
1.  
Pump pressure too low.  
a. See 5.4  
2.  
Low feed water temperature.  
a. Verify product production with Tech Note 121 Temperature Correction.  
b. Install or adjust temperature blending valve.  
3.  
Membrane fouled or scaled.  
a. Review trend analysis logs for flow.  
b. Clean membrane with both high and low pH cleaners.  
c. If flow is not recoverable, replace membrane.  
12552e - 27 Feb 04  
5- 7  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.  
5.  
Center tube of ROMA collapsed from operating at high temperatures.  
a. Correct temperature problem.  
b. Replace membrane.  
Check softener pretreatment for brackish (high salt) water.  
a. Check inlet water for high conductivity.  
b. Check water softener for proper function.  
c. Regenerate softener.  
d. Check supply water for hardness and conductivity.  
6.  
Feed water fouling membrane (change in feed water).  
a. Check for change in feed water quality.  
b. Check with city water department.  
5.8  
Product Flow Too High  
1.  
Rejection Satisfactory  
A. Pump pressure too high.  
a. Adjust pressure regulator setting to specifications.  
B. Feedwater temperature too high.  
a. Adjust temperature blending valve.  
b. Correct plumbing error that allows the cold water to be heated.  
2.  
Rejection Unsatisfactory  
A. Waste flow too low.  
a. Check waste flow rate.  
b. Check for blockage or kink in waste line.  
c. Quick disconnect not connected completely.  
B. High pH and chlorines in feed water.  
a. Check feedwater for higher than 8.5 and the presence of chloramines.  
b. Check pH reduction equipment if applicable.  
c. Contact GE Osmonics.  
C. ROMA internal seals damaged, dirty, or dislodged.  
a. Disassemble ROMA, clean and inspect “O” rings and seals.  
D. Membrane hydrolyzed (exposed to oxidant).  
a. Verify condition of pretreatment equipment.  
E. Membrane damaged (or ruptured).  
a. Check for high product pressure.  
b. Check for product line being back pressured.  
F. Normal membrane rejection decline (over time).  
a. Perform AAMI water lab analysis.  
b. Replace membrane if indicated by analysis.  
12552e - 27 Feb 04  
5- 8  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
5.9  
Membrane Product Pressure Too Low  
1.  
Product flow rate below specifications for temperature and pressure.  
a. Clean or replace membrane.  
2.  
Flow demand exceeds flow capacity of membrane.  
a. Correct condition demanding excessive flows.  
b. Increase temperature of feed water and/or pump pressure (within operating  
limits). Continuously monitor temperature.  
5.10 Leaks  
1.  
General information: Any leak from any part of the system should be corrected  
as soon as possible. Left unattended, leaks can cause secondary damage  
necessitating expensive repairs.  
2.  
Refer to the Maintenance Procedures for Leak Repair.  
5.11 Low Percent Rejection  
The overall membrane performance can be affected by a number of factors. Poor  
rejection performance, as well as high TDS, can result from fouled membranes, low  
pump pressure, inadequate waste flow, or feed water problems. An apparent  
performance problem could be caused by a defective sensor or circuit board, and not be  
a membrane issue at all. The recommended first step for any ‘membrane performance’  
question is to confirm the rejection percentage and product TDS with a hand-held  
RO/TDS meter. If the product water quality is poor, investigate the membrane.  
However, if the water quality tests ‘normal’, investigate the monitoring circuits.  
1.  
Disinfectant or cleaner left in RO unit.  
a. Check for presence of disinfectant or cleaner, rinse as necessary.  
2.  
Waste flow too low.  
a. Pump not operating correctly, see 5.4.  
b. Check waste flow control.  
c. Check for obstruction or kink in waste line.  
3.  
Pump pressure too low.  
a. Adjust pressure regulator.  
b. Adjust pump settings.  
c. Pump not operating correctly, see 5.4.  
4.  
5.  
Product flow too high.  
a. See 5.8  
Membrane scaled or fouled.  
a. Verify condition of pretreatment equipment.  
b. See membrane cleaning procedure in this manual.  
12552e - 27 Feb 04  
5- 9  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
6.  
Erroneous meter readout.  
a. Verify water quality with independent meter and check conductivity probe  
connection.  
b. Check all quick-disconnect connections.  
c. Replace inlet water sensor probe.  
d. Replace product water sensor probe (percent rejection).  
e. Replace monitor circuit board.  
7.  
8.  
Membrane failure.  
a. Replace membrane.  
Change in feed water TDS.  
a. Verify feed and product water quality with independent meter.  
b. Softener problem, rinsing out high levels of salt.  
9.  
Normal membrane rejection decline (over time).  
a. Perform AAMI water lab analysis.  
b. Replace membrane if indicated by analysis.  
5.12 Rejection Display Fluctuates  
1.  
Display cycles (fluctuates) up and down.  
a. Check for defective product recirculation check valve and replace if necessary.  
5.13 TDS Value High  
The overall membrane performance can be affected by a number of factors. Poor  
rejection performance, as well as high TDS, can result from fouled membranes, low  
pump pressure, inadequate waste flow, or feed water problems. An apparent  
performance problem could be caused by a defective check valve, sensor or circuit  
board, and not be a membrane issue at all. The recommended first step for any  
‘membrane performance’ question is to confirm the rejection percentage and product  
TDS with a hand-held RO/TDS meter. If the product water quality is poor, investigate  
the membrane. However, if the water quality tests ‘normal’, investigate the monitoring  
circuits.  
1.  
Disinfectant or cleaner left in RO unit.  
a. Check for presence of disinfectant or cleaner, rinse as necessary.  
2.  
Air bubble trapped on sensor probe.  
a. With RO operating, open and close the product valve to dislodge air.  
b. With RO operating, rotate the Dialyze/Flush valve back and forth to dislodge  
air.  
3.  
Waste flow too low.  
a. Measure and compare to original data. Correct as necessary.  
12552e - 27 Feb 04  
5- 10  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
4.  
5.  
6.  
Waste pressure too low.  
a. Adjust pump settings.  
Product flow too high.  
a. See 5.8  
Membrane scaled or fouled.  
a. Verify condition of pretreatment equipment.  
b. See membrane cleaning procedure in this manual.  
7.  
Erroneous meter readout.  
a. Verify water quality with independent meter.  
b. See 5.12.7  
c. TDS water probe wire disconnected or loose.  
d. Replace TDS water sensor probe.  
e. Replace thermistor.  
f. Replace monitor circuit board.  
8.  
9.  
Membrane failure.  
a. Replace membrane.  
Change in feed water TDS.  
a. Verify feed and product water quality with independent meter.  
b. Communicate with municipal water plant.  
c. Softener problem, rinsing out high levels of salt.  
10.  
Normal membrane rejection decline (over time).  
a. Perform AAMI water lab analysis.  
b. Replace membrane if indicated by analysis.  
5.14 RO Fails to Draw Disinfectant  
1.  
Pre-filter bowl leaking on vacuum.  
a. Tighten bowl  
b. Inspect and clean pre-filter bowl seal / replace seal.  
c. Replace filter bowl  
2.  
3.  
Disinfect line kinked or plugged.  
a. Clear disinfect line.  
Air leak on disinfect draw line.  
a. Check that quick disconnect fitting is connected securely.  
b. Check tube for cracks or holes.  
c. Replace quick disconnect fitting.  
4.  
Motor fails to turn on.  
a. See 5.1  
12552e - 27 Feb 04  
5- 11  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
5.15 RO Pump Leaks  
1.  
2.  
Tubing connection leaking.  
a. Tighten or reconnect tubing fitting.  
Mechanical seal failure (leaks out vent hole).  
a. Replace pump.  
12552e - 27 Feb 04  
5- 12  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
NOTES:  
12552e - 27 Feb 04  
5- 13  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e - 27 Feb 04  
5- 14  
Troubleshooting  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
F-801 Internal RO System  
CHAPTER SIX:  
DRAWINGS  
12552e – 27 Feb 04  
6 - 1  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e – 27 Feb 04  
6 - 2  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
REMOTE MONITOR  
MODULE  
PANEL  
F-801 RO and Cart Assembly  
(Component Layout)  
12552e – 27 Feb 04  
6 - 3  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
PREFILTER  
CARBON TANK  
SHORT BACK  
PREFILTER  
CARBON TANK  
SOFTENER  
PREFILTER  
CARBON TANKS  
* A three tank option is available, please consult factory.  
F-801 RO  
(Typical Pretreatment Configurations)  
12552e – 27 Feb 04  
6 - 4  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
Assembly Instructions  
1.  
2.  
Remove plugs from the tank head.  
Install the input elbow, output nipple, and the sample valve in locations shown.  
Use Teflon tape and thread in approximately 3-5 turns.  
OUTPUT  
INPUT  
TANK HEAD  
40480  
SAMPLE VALVE  
40291  
OUTPUT NIPPLE  
40268  
INPUT ELBOW  
40268A  
F-801 RO  
(GAC Tank Fitting Assembly)  
12552e – 27 Feb 04  
6 - 5  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
40571  
FEED WATER FITTING  
40569  
ELBOW  
ADAPTER  
40564  
40572  
PRODUCT WATER  
FITTING  
40570  
DUAL PORT  
40599A  
12104  
2.5" X 40" ROMA  
12093  
SINGLE PORT  
12103  
F-801 RO  
(ROMA Part Description)  
12552e – 27 Feb 04  
6 - 6  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
60158  
40704  
40738  
IN  
OUT  
*40095  
*60370  
*60240  
*40605  
*40095  
F-801 RO  
(Pre-filter Assembly)  
12552e – 27 Feb 04  
6 - 7  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
MANIFOLD  
ASSEMBLY  
14731  
MOTOR  
30014  
PUMP ASSEMBLY  
12168−1  
INPUT  
ASSEMBLY  
12285−1  
F-801 RO  
(Plumbing Module, Main Components)  
12552e – 27 Feb 04  
6 - 8  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
60360  
30014  
SEE  
DETAIL ’B’  
12285−1  
(SEE DETAIL ’A’)  
50003  
12922  
12168−1  
40568  
40569  
1234489  
40676  
12131  
50012  
DETAIL ’A’  
DETAIL ’B’  
F-801 RO  
(Plumbing Module, Component Details)  
12552e – 27 Feb 04  
6 - 9  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
40576  
(2X)  
40564  
PART OF ITEM  
40947  
40947  
40948  
40561  
PART OF ITEM  
40948  
F-801 RO  
(Plumbing Module, Component Details)  
12552e – 27 Feb 04  
6 - 10  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
40564  
40561  
12922  
40742  
40677  
14731  
60002  
60041  
40569  
41614  
12130  
50112  
F-801 RO  
(Plumbing Module, Component Details)  
12552e – 27 Feb 04  
6 - 11  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
F-801 RO  
(Plumbing Module, Tubing)  
12552e – 27 Feb 04  
6 - 12  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
A (12")  
B (8")  
C (5 3/4")  
B (5 3/4")  
B (3 1/2")  
See Detail ’A’  
on next sheet  
See Detail ’B’  
on next sheet  
A (5")  
A (10")  
D (2")  
C (15 1/2")  
B (12 1/4")  
D (10")  
A (10")  
D (45")  
D (3 3/4")  
LEGEND  
Item Letter  
Description  
3/8” Tubing, Translucent (Natural)  
3/8” Tubing, White  
1/4" Tubing, Translucent (Natural)  
1/8” Tubing, White  
Part No.  
60020  
60032  
60184  
60298  
A
B
C
D
F-801 RO  
(Plumbing Module, Tubing)  
12552e – 27 Feb 04  
6 - 13  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
41036  
40671  
40254  
40143  
40077  
40602  
DETAIL ’A’  
40564  
12288  
DETAIL ’B’  
F-801 RO  
(Plumbing Module, Component Details)  
12552e – 27 Feb 04  
6 - 14  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
60307  
12110  
40332  
40584  
20178  
40602  
12131  
SEE DETAIL ’A’  
12201  
40570  
60051  
40569  
40564  
40218  
40571  
12131  
PLUS SIDE  
DETAIL ’A’  
Note:  
The ‘plus side’ of the check valve (p/n 40584) should face towards the small manifold  
(p/n 12201).  
F-801 RO  
(Manifold Assembly)  
12552e – 27 Feb 04  
6 - 15  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
20177  
12359  
12329  
12213  
30026  
12310  
12312  
12324  
12315  
F-801 RO  
(Monitor Module, Front Section)  
12552e – 27 Feb 04  
6 - 16  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
12776  
30010  
FUSE INSIDE  
(20012)  
60002  
20203  
60041  
12311  
F-801 RO  
(Monitor Module, Back Section)  
12552e – 27 Feb 04  
6 - 17  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
NOTE:  
ALL TUBING IS 3/8" I.D. BRAIDED HOSE  
UNLESS OTHERWISE NOTED.  
RO MEMBRANE PERMEATE TUBING  
(1/4" OD POLYETHYLENE TUBE)  
(65" LENGTH OF 60184)  
PREFILTER HOUSING  
ASSEMBLY  
ADAPTER, 3/8" HOSE  
TO 3/8" STEM  
(40672)  
(12197)  
GREEN MALE  
Q.D. (12886)  
GREEN FEMALE  
Q.D. (12734)  
RO PUMP DISCHARGE HOSE  
TO RO HOUSING INLET  
(12208)  
MEMBRANE HOUSING  
ASSEMBLY  
(12173)  
PREFILTER FEEDWATER  
HOSE (CONNECT TO CARBON  
TANK) (12388)  
FEED  
F−801 RO SYSTEM  
FEEDWATER HOSE  
RO MEMBRANE  
HOUSING  
F−801 RO SYSTEM  
FLUID CHASSIS  
(12174)  
SUPPORT  
SCREWS  
RO MEMBRANE WASTE HOSE  
(12374)  
PRODUCT  
WASTE  
PRODUCT−TO−USE HOSE  
2 PIECE (12205)  
RED MALE Q.D.  
(12731)  
WHITE FEMALE Q.D. (40693)  
WHITE MALE Q.D. (40717)  
VALVE (40316)  
RED FEMALE  
Q.D. (12733)  
HOSE ADAPTER (40581)  
WASTE−TO−DRAIN HOSE  
2 PIECE (12207)  
FLUSH HOSE (12393)  
WHITE FEMALE Q.D. (40762)  
F-801 RO  
(General Hose Connections)  
12552e – 27 Feb 04  
6 - 18  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
GREEN QUICK DISCONNECT  
RED QUICK DISCONNECT  
F−801 WATER SUPPLY HOSE. REMOVE  
FROM PREFILTER OUTLET PORT AND  
CONNECT TO DISINFECT TANK SUCTION  
PORT FITTING.  
F−801 WASTE−TO−DRAIN.  
DISCONNECT QUICK−CONNECT  
FITTING FROM REST OF HOSE,  
AND CONNECT TO DISINFECT  
F−801 PRODUCT−TO−USE HOSE. DISCONNECT  
QUICK−CONNECT FITTING FROM REST OF  
HOSE, AND CONNECT TO DISINFECT TANK.  
F-801 RO  
(Disinfect / Cleaning Connections)  
12552e – 27 Feb 04  
6 - 19  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
F-801 RO  
(High Voltage Electrical, 60Hz)  
12552e – 27 Feb 04  
6 - 20  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
D R E  
D R E  
N E E G R  
W O L L Y E  
N E E G R  
F-801 RO  
(Low Voltage Electrical)  
12552e – 27 Feb 04  
6 - 21  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
F-801 RO  
(Plumbing Schematic)  
12552e – 27 Feb 04  
6 - 22  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
TOOLS REQUIRED:  
1. FLAT BLADE SCREWDRIVER.  
2. 3/4" WRENCH OR EQUIVALENT.  
3. TOWEL.  
PROCEDURE:  
1. TURN OFF POWER AND WATER TO UNIT.  
2. OPEN FRONT COVER AND EXPOSE PUMPHEAD.  
3. REMOVE ACORN NUT. NOTE: DO NOT LOSE WASHER.  
4. USE SCREWDRIVER AND TURN SCREW AS SHOWN. CLOCKWISE TO INCREASE PRESSURE, COUNTER  
CLOCKWISE TO DECREASE PRESSURE. EACH 90° ROTATION WILL CHANGE PRESSURE ABOUT 25 PSI.  
5. REPLACE WASHER (IF IT CAME OFF) AND ACORN NUT.  
6. TURN ON WATER AND POWER AND THEN CHECK PRESSURE READING. REPEAT AS NECESSARY.  
PRESSURE ADJ. SCREW  
DECREASE  
INCREASE  
WASHER  
ACORN NUT  
PUMP  
F-801 RO  
(Pump Pressure Adjustment Procedure)  
12552e – 27 Feb 04  
6 - 23  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
“Push-In” Fitting Instructions  
12552e – 27 Feb 04  
6 - 24  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
NOTES:  
12552e – 27 Feb 04  
6 - 25  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e – 27 Feb 04  
6 - 26  
Drawings  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
F-801 Internal RO System  
CHAPTER SEVEN:  
SPARE PARTS LIST  
12552e – 27 Feb 04  
7 - 1  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e – 27 Feb 04  
7 - 2  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
F-801 Internal RO System  
SPARE PARTS LIST  
NOTE: Part Numbers are subject to change and should be verified when ordering  
spare parts.  
CATALOG NUMBER  
DESCRIPTION  
PARTS  
12093  
12103  
12104  
12106  
12128  
12130  
12131  
12163  
12168-1  
12173  
12197  
12211  
12213  
12285-1  
12288  
12310  
12311  
12312  
12315  
12318  
12320  
12331  
12334  
12337  
12352  
12355  
12357  
12359  
12496  
12497  
12498  
12499  
12885  
12888  
Housing for ROMA  
End cap, Waste  
End cap, Input / Product  
Retaining Pin  
Pre-filter Housing Assembly  
Flush Valve Bracket  
Sensor Probes, Total Dissolved Solids, Feed, Product  
Connector Rod  
Pump assembly  
ROMA, 40”, TF (complete assembly)  
Pre-filter assembly  
Disinfect tank assembly  
8-32 Painted Screw  
Manifold assembly, Input  
Flow control orifice  
Printed Circuit Board, % Rejection  
Printed Circuit Board, Conductivity  
Printed Circuit Board, Display  
Power Supply, 115~  
ROMA mounting strap  
ROMA / Monitor mounting bracket  
Stand, 2 tank  
Spacer, undercarriage  
Leg, Left  
Pre-treat strap, V-style  
Spacer, aluminum  
Leg, Right  
Lens, Red, for Display PCB  
Cabinet attaching kit  
Carbon tank fitting kit  
Hardware kit  
ROMA fitting kit  
Stand, 2 tank handle assembly  
Hose Kit (For all hose connections)  
12552e – 27 Feb 04  
7 - 3  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
CATALOG NUMBER  
DESCRIPTION  
PARTS (cont.)  
12922  
14398  
14399  
14665  
14731  
20012  
20177  
20178  
20203  
30010  
30014  
30026  
40077  
40080  
40092  
40093  
40095  
40165  
40291  
40332  
40480  
40571  
40572  
40584  
40599A  
40602  
40676  
40694  
40696  
40704  
40741  
40947  
40948  
41555  
41614  
50003  
50012  
50017  
50041  
60002  
10-32 Painted Screw  
Pre-treat strap, A-style  
Pre-treat strap, B-style  
Manual, F-801 Internal RO System  
Manifold assembly, Main  
Fuse, 10 Amp, Slo Blow  
Lens cover, for LED  
Thermistor  
Switch, RO Pump Power  
Fuse Holder, Panel mount  
Motor, 1/3 HP, 110/230 V, 50/60 Hz  
Selector knob  
Tee, PVC, 1/4" fnpt  
Nipple, PVC, 3/8 x short  
O-ring, large  
O-ring, small  
Washer, rubber, for Female Garden Hose Fittings  
Pump, 100 GPH, SS, W/Regulator  
Sample Valve  
O-ring, for product recirculation check valve  
GAC tank head  
Lockclip, 3/8” SF  
Lockclip, 1/4" SF  
Product recirculation check valve  
Membrane, TF, 40”  
Reducer bushing, 1/4 mnpt x 1/8 fnpt  
Nipple, PVC, 1/4 mnpt x close  
O-ring, Pre-filter (9 7/8”)  
Filter, 5 Micron, 9 7/8"  
Pre-filter mounting bracket  
Gauge, 100 PSI, pre-filter  
Gauge, 100 PSI  
Gauge, 300 PSI  
Hose Clamp, 5/8”  
Valve, ball, 1/4 fnpt  
Clamp, Pump to Motor  
Rubber foot  
Caster, Dual Wheel  
Pump-to-Motor Coupling  
Clamp, strain, 3/4"  
12552e – 27 Feb 04  
7 - 4  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
CATALOG NUMBER  
DESCRIPTION  
PARTS (cont.)  
60041  
60278  
60307  
60360  
60369  
1234489  
Clamp, strain, 7/8”  
Grommet, 7/8x5/8x16GA  
Terminal, QD, 18-22GA, 1/4" tab female  
Motor mount  
Cotter Pin, for retaining pin  
Regulator, Product Pressure Adjustment  
FITTINGS  
12612  
12613  
12731  
12732  
12734  
12886  
40143  
40218  
40254  
40268  
40268A  
40561  
40564  
40568  
40569  
40570  
40576  
40605  
40671  
40672  
40677  
40693  
40716  
40717  
40742  
40743  
41036  
41083  
41566  
Quick-disconnect (Male) for Disinfect container, Red  
Quick-disconnect (Male) for Disinfect container, Green  
Quick-disconnect (Male) for Waste hose, Red  
Quick-disconnect (Female) for Waste hose, Red  
Quick-disconnect (Female) for Feed hose, Green  
Quick-disconnect (Male) for Pre-filter outlet, Green  
Tee, 1/8” tube  
Elbow, 1/4" tube x 1/8 mnpt  
Straight, 1/8” tube x 1/8” mnpt  
Straight, Garden Hose Thread, male  
Elbow, Garden Hose Thread, male  
Union, speed-fit bulkhead, 3/8  
Elbow, speed-fit, 3/8 x 3/8 stem  
Elbow, speed-fit, 3/8 x 3/8 mnpt, swivel  
Elbow, speed-fit, 3/8 x 1/4 mnpt, swivel  
Elbow, speed-fit, 1/4 x 1/4 mnpt, swivel  
Straight, 1/8” tube x fnpt  
Adapter, Garden Hose Thread x 3/8 barb  
Straight, speed-fit, 3/8 x 1/4 mnpt  
Adapter, speed-fit, 3/8 x 3/8 barb  
Union, speed-fit bulkhead, 1/4  
Quick-disconnect (Female) for Product hose, White  
Quick-disconnect (Female) for Disinfect container, White  
Quick-disconnect (Male) for product hose, White  
Elbow, speed-fit, 1/4 x 1/4 stem  
Elbow, PVC, 3/8 fnpt  
Elbow, speed-fit union, 3/8  
Straight, speed-fit, 3/8 x 1/4 fnpt  
Elbow, speed-fit, 1/4 x 3/8 stem  
12552e – 27 Feb 04  
7 - 5  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
CLEANERS  
14962  
Membrane Cleaner, BIOSAN, 1 gal., liquid  
1228653  
1228791  
Cleaner, LIQUICLEAN 103™,1gal., Low pH, liquid  
Cleaner, LIQUICLEAN 511™,1gal., High pH, liquid  
TESTS  
11656  
14604  
50033  
70001  
Total Chlorine Test Kit  
Silt Density Index (SDI) Test Kit  
Hardness Test Kit  
Water Test Kit, AAMI Standards  
OPTIONAL  
40695  
40698  
Filter, 1 Micron, 9 7/8"  
Filter, Carbon, 9 7/8"  
12552e – 27 Feb 04  
7 - 6  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
NOTES:  
12552e – 27 Feb 04  
7 - 7  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
F-801 Internal RO System by GE Osmonics  
This page intentionally left blank.  
12552e – 27 Feb 04  
7 - 8  
Spare Parts List  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Call (253) 395-2200 for additional information, (800) 633-3080 in the U.S., or visit www.osmonics.com  
Corporate Headquarters  
5951 Clearwater Drive  
Minnetonka, MN  
55343-8995  
Kent Operations  
7848 South 202nd ST  
Kent, WA  
98032-1345  
USA  
USA  
(952) 933 2277 Phone  
(952) 933-0141 Fax  
(253) 395-2200 Phone  
(253) 395-2363 Fax  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Electro Voice Speaker System PRO 8A User Manual
Fisher Price Motorized Toy Car B1476 User Manual
Fisher Price Video Games L6367 User Manual
FoodSaver Food Saver V2490BC User Manual
Fusion Stereo Amplifier PP AM100010 User Manual
Gateway Computer Monitor VX730 User Manual
GE Monogram Refrigerator ZISP480DTSS User Manual
GE Monogram Ventilation Hood ZV755SPSS User Manual
GE Washer Dryer DSKS433ET User Manual
Gianni Industries Door L 10010N User Manual